- Industrial & lab equipment
- Measuring, testing & control
- Hanna Instruments
- HI 2300
- Instruction manual
Hanna Instruments HI 2300 Instruction manual
Add to my manuals
20 Pages
Hanna Instruments HI 2300 is a bench microprocessor-based meter designed to measure conductivity, TDS, NaCl, and temperature. It offers a variety of features for accurate and convenient measurements, including autoranging, temperature compensation, calibration, data storage, and computer connectivity. With its high accuracy and resolution, HI 2300 is ideal for various applications in laboratories, water treatment plants, and industrial settings.
advertisement
Instruction Manual
HI 2300
Auto-ranging
Microprocessor
EC/TDS/NaCl/ºC
Bench Meter
ALT
GLP
SETUP
CFM
Functions
RANGE
CFM
ATC
CAL
/
LOCK
Start Calibration
Lock Current Range
Change V
Confirm V alue
ATC
TC
ALT Functions
CAL
T
SETUP
Calibration
Enter Setup Mode
TC
GLP
View Calibration
Display TC Factor
RANGE
LOCK
C w w w. h a n n a i n s t . c o m
1
Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using the instrument.
This manual will provide you with the necessary information for correct use of the instrument, as well as a precise idea of its versatility.
If you need additional technical information, do not hesitate to e-mail us at [email protected] or turn to the back cover for our worldwide contact list.
This instrument is in compliance with directives.
W A R R A N T Y
HI 2300 is guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Electrodes and probes are guaranteed for six months.
This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods
Authorization number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
TABLE OF CONTENTS
WARRANTY ..................................................................................................... 2
PRELIMINARY EXAMINATION .............................................................................. 3
GENERAL DESCRIPTION .................................................................................... 3
FUNCTIONAL DESCRIPTION ................................................................................ 4
SPECIFICATIONS ............................................................................................... 5
OPERATIONAL GUIDE ........................................................................................ 7
AUTORANGING ................................................................................................ 8
TEMPERATURE COMPENSATION ......................................................................... 8
EC/TDS CALIBRATION ....................................................................................... 9
NaCl CALIBRATION ......................................................................................... 10
TEMPERATURE CALIBRATION (for technical personnel only) .................................. 12
TEMPERATURE ADJUSTMENT .......................................................................... 12
CONDUCTIVITY VERSUS TEMPERATURE CHART ................................................. 13
SETUP .......................................................................................................... 14
GOOD LABORATORY PRACTICE ........................................................................ 15
DATA TRANSFER TO PC ................................................................................... 17
PROBE MAINTENANCE .................................................................................... 18
ACCESSORIES ................................................................................................ 18
2
P R E L I M I N A R Y E X A M I N A T I O N
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer or the nearest Hanna Customer Service Center.
Each instrument is supplied with:
• HI 76310 Conductivity / TDS probe
• 12V Power Adapter
• Instruction Manual
Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in the original packing with the supplied accessories.
G E N E R A L D E S C R I P T I O N
The HANNA HI 2300 is a bench microprocessor-based Conductivity/TDS/
NaCl/Temperature meter.
The autoranging feature of the EC and TDS ranges automatically sets the instrument to the scale with the highest possible resolution.
The conductivity measurements are manually or automatically compensated for temperature effect, with the temperature sensor inside the conductivity probe. It is also possible to disable the temperature compensation and measure the actual conductivity.
The temperature coefficient is user selectable.
The instrument is equipped with a stability indicator. With this feature the user will always know when to record a measurement.
HI 2300 includes also GLP capability and data transfer to a computer through a RS232 port.
In addition, the meter allows the user to enter an ID code to uniquely identify the instrument.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
3
F U N C T I O N A L D E S C R I P T I O N
Front Panel Primary LCD
Secondary LCD
1
CAL
CFM g / l
NaCl ppm
BUF
%
C
2
3
4
5
6
7
CAL
CALT
CFM
GLP SETUP
ALT
ATC
TC
Functions
RANGE
CAL
Select Measurement
Range
Start Calibration
LOCK
/
CFM
ATC
Lock Current Range
Change Value
Confirm Value
Temperature
Compensation
ALT Functions
CALT
SETUP
GLP
TC
Start Temperature
Calibration
Enter Setup Mode
View Calibration
Data
Display TC Factor
RANGE
LOCK
HI 2300
Microprocessor Conductivity Meter
10
8
9
Rear Panel
POWER
12VDC
PROBE
RS232
11 12 13
1) Liquid Crystal Display (LCD).
2) CAL key, to enter/exit calibration mode.
CALT key (alternate function), to enter temperature calibration mode.
3) CFM key, to confirm different values.
4) key, to manually decrease temperature value or other parameters.
SETUP key (alternate function), to enter/exit SETUP mode.
5) key, to manually increase temperature value or other parameters.
GLP key (alternate function), to display Good Laboratory Practice information.
6) ATC key, to select temperature compensation mode.
TC key (alternate function), to view the temperature coefficient value.
7) ALT key, to select alternate key function.
8) LOCK key, to freeze current range on the LCD.
9) RANGE key, to select the measurement unit or to switch the focused data.
10) ON/OFF switch.
11) Power supply socket.
12) Probe connector.
13) RS232 serial communication connector.
4
S P E C I F I C A T I O N S
R
@
C A
E S
R
O
A
A C C U
2
L I
0
B
L
R
N G
°
U
E
T I O N
R A C Y
A
C /
T I
6 8
O N
° F
T
T
T
T
T
N
T
N
T
N
N e e e e
E
E
E
E
D
D
D a
C m a
C m a m a
C
C m
S
S
C l
C l
S
C l
C l p p p p u p
0
3
3
.
0
0
0
.
0
0
0 t o t o t o
2 9 .
9 9
2 9 9 .
9
2 9 9 9
µ S
µ S
/ c m
/ c m t o
3
3
.
0
0
.
0
0
5 0 0 .
0 t o t o
2
2
9
0
.
0
9
.
9
0 m S / c m (
µ S / c m m S / c m m S / c m a tc u a l E C ) *
0
1
.
5
0
.
0
0
1 5 .
0 t o t o t o
1
1
4
4
.
9
9
.
9
9
1 0 0 .
0 m m g g
/
/
1 5 0
1 .
5 0 t o t o
1 4 9 9
1 4 .
m
9 9 g g
/ L
/ L g / L
L
L
(
( p
( p p p m ) m )
(
( p p p p m )
)t p p )t u p w t o ti h
4 0 0 .
0
0 .
8 0 g / L ( a c o n v e r is o tc u a l n f a
T D tc o r
S ) *
0 .
0 t o 4 0 0 .
0 %
– 9 .
9 t o 1 2 0 .
0 ° C
0 .
0 1 µ S / c m
0 .
1 µ S / c m
0 .
1 µ S / c m
0 1 m S / c m
0 .
1 m S / c m
0 .
0 1 m g / L
0 .
1
1 m g m g / L
/ L
0 .
0 1
0 .
1 g g
/ L
/ L
0 .
1 %
0 .
1 ° C
± 1 % o f r e a d i n g
1 d i
± ( 0 .
0 5 g i )t
µ S / c m o r
± 1 % o f r e a d i n g ± ( 0 .
0 3
1 d i g i )t m g / L o r
± 1 % o f r e a d i n g
± 0 .
4 ° C
5 .
0 0
A u t o m a
, m e m o
8 4 .
0 , ir z e d
1 4 1 3 v a
µ
S l u e
/ c m
:s
1 2 .
8 8 , it ,c 1 p o i n t
8 0 .
0 , w
1 1 1 .
8 ti h 6 m S / c m
1 p o i n ,t w ti h H I s o l u
7 0 3 7 it o n c a il b r a it o n
2 p o i n ,t a t 0
1 a n d
2 2
5 0
° F )
° C ( 3 2 a n d
* Actual conductivity (or TDS) is the conductivity (or TDS) value without temperature compensation.
5
SPECIFICATIONS (cont.) c
T e m p e o m p e n r a t u r e s a it o n
T e m p e r a t u r e c o e f if ic e n t
T D S c o n v e r is o n f a tc o r
P r o b e i( n lc u d e d )
C o m p u t e r I n t e fr a c e
A u t o o ff
P o w e r s u p p l y
D i m e n is o n s
W e i g h t
E n v ri o n m e n t
W a rr a n t y
A u t o m a t ci o r
( 3 2 m a n u t o 1 4 a
0
,l 0
º F ) t o 6 0 º C
S e l e tc a b l e
( E C rf o m a n d
0 .
0 0
T D S t o 6 .
0 0 o n l y )
% / º C
S e l e tc a b l e
( d e f a u tl rf o m v a
0 .
l u e :
4 0 t o
0 .
5 0 )
0 .
8 0
H I 7 6 3 1 0 4 t e m ir n g P p e r a t u r e l a it n u m , s e n s o r i n t e r n a l
R S 2 3 2 o p t o si o l a t e d a tf e r 5 m
( c a n i n b e u t e s d o f n o si a b l e d ) n u s e
1 2 V d c i( n p o w e r a d a p t e r lc u d e d )
2 4 0 x 1 8 2 x 7 4 m m ( 9 .
4 x 7 .
1 x 2 .
9 ” )
1 .
1 K g ( 2 .
4 l b )
0 m a x
–
R H
5 0 º C
9 5 %
( 3 2 n o n
1 2 2 c o n d e n
º F ) is n g
2 y e a sr
6
OPERATIONAL GUIDE
CONNECTIONS
Plug the 12 VDC adapter into the power supply socket.
Note: This instrument use non volatile memory to retain the calibration parameters and all the other settings even when unplugged.
Connect the EC/TDS probe to the 7-pin connector. Tighten the threaded ring.
Make sure the probe sleeve is properly inserted, as shown below:
POWER
12VDC
PROBE
RS232
INSTRUMENT START-UP
• Turn the instrument on by pressing the ON/OFF switch.
CAL
CFM g / l
NaCl ppm
BUF
%
C
• All LCD tags are displayed and a beep is generated while the instrument performs self test.
TAKING MEASUREMENTS
Immerse the probe into the solution to be tested. The sleeve holes must be completely submerged. Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.
If needed, press RANGE repeatedly until the desired range (EC, TDS, NaCl) is selected on the LCD (displaying µS[mS], ppm[g/L] or %).
C
RANGE
Allow for the reading to stabilize. The upper LCD displays the measurement in the selected range, while the temperature is displayed on the lower LCD.
7
Notes:
• If the meter displays “----”, the reading is out of range.
• If the stability indicator “~” blinks, the reading is unstable.
• Make sure the meter is calibrated before taking measurements.
• If measurements are taken successively in different samples, for accurate readings it is recommended to rinse the probe thoroughly with deionized water before immersing it into the samples.
•
• TDS reading is obtained by multiplying the EC reading by the TDS factor, which has a default value of 0.50. It is possible to change the
TDS factor in the 0.40 to 0.80 range by entering setup mode and selecting the “tdS” item (see SETUP for details).
When the use of an alternate function (SETUP, TC, GLP or CALT) is requested, press and hold the ALT key first, and then the second key.
A U T O R A N G I N G
The EC and TDS scales are autoranging. The meter automatically sets the scale with the highest possible resolution.
By pressing LOCK, the autoranging feature is disabled and the current range is frozen on the LCD. “1” tag starts blinking. To restore the autoranging option press LOCK again.
LOCK
Note: Autoranging is automatically restored if the range is changed, if the setup or calibration modes are entered, or if the meter is turned off and back on again.
T E M P E R A T U R E C O M P E N S A T I O N
Three options of compensating temperature are available:
Automatic (Atc): The EC probe has a built-in temperature sensor; the temperature value is used to automatically compensate the EC/TDS reading
(from –9.9 °C to 120.0 °C), using the selected reference temperature.
Manual (Mtc): The temperature value, shown on the secondary LCD, can be manually set with the ARROW keys. The compensation is referenced at the selected reference temperature. The “°C” tag blinks when this option is active.
No Compensation (notc): The temperature value shown on the secondary LCD is not taken into account. The reading displayed on the primary LCD is the actual
EC or TDS value. The “°C” tag blinks when this option is active.
Note: The default compensation mode is ATC.
8
To select the desired option press ATC until the option is displayed on the LCD.
If temperature compensation is active, measurements are compensated using the temperature coefficient (default value
1.90 %/°C).
ATC
To change the temperature coefficient, enter the setup mode and select the “tc” item (See SETUP for details, page 14). The current temperature coefficient can be quickly viewed by pressing ALT+TC keys. The value is briefly displayed on the secondary LCD.
ALT TC
• If the temperature reading is out of –9.9 °C - 120.0 °C interval and Atc or
Mtc option is selected, the temperature limit range value will be displayed, together with the “°C” tag blinking and the instrument will do no temperature compensation.
E C / T D S C A L I B R A T I O N
EC calibration is a one-point procedure. Selectable calibration points are:
0.00 µS, 84.0 µS, 1413 µS, 5.00 mS, 12.88 mS, 80.0 mS, 111.8 mS.
To enter EC calibration, select the EC range and press CAL.
CAL
Note: TDS reading is automatically derived from the EC reading and no specific calibration for TDS is needed. Pressing CAL when TDS range is selected has no effect.
Rinse the probe with calibration solution or deionized water.
Immerse the probe into the solution. The sleeve holes must be completely submerged. Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.
For zero calibration, just leave the dry probe in the air.
The “BUF” and “CAL” tags are displayed. The primary LCD displays the uncalibrated EC reading. The secondary LCD displays the buffer value. The
“~” tag blinks.
CAL
BUF
9
Select the desired value with the ARROW keys, if necessary.
OR
When the reading is stable, “CFM” tag starts blinking on the LCD, asking for confirmation. Press CFM to confirm calibration.
CAL
CFM
BUF
CFM
The instrument displays the “Stor Good” message and returns to measurement mode.
Notes:
• If the uncalibrated reading is too far from the expected value or the temperature value is out of 0 – 60 ºC range, calibration is not recognized.
The “CFM” tag does not appear; the “~” and “BUF” tags blink to signal wrong or contaminated calibration solution. If the temperature value is out of 0 – 60 ºC range, the “ºC” tag will also be displayed blinking on the LCD.
•
For best results choose an EC buffer value close to the sample to be measured.
• In order to minimize any EMC interferences, use plastic beakers.
• The meter uses 1.90 %/°C compensation factor during calibration. If the setup item “tc” has been set to a different value, when exiting calibration mode, the displayed value on the upper LCD might be different from the nominal buffer value.
• It is possible to set the cell constant value directly, without following the calibration procedure. To set the cell constant enter the setup mode and select “CEL” (see SETUP for details).
N a C l C A L I B R A T I O N
NaCl calibration is a one-point procedure at 100.0% NaCl. Use the
HI 7037L calibration solution (sea water solution) as a 100% NaCl standard solution.
To enter NaCl calibration select the NaCl range and press CAL.
CAL
Rinse the probe with some of the calibration solution or deionized water. Immerse the probe into HI 7037L solution. The sleeve holes must be completely submerged.
Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.
10
The “BUF” and “CAL” tags are displayed. The primary LCD displays the uncalibrated NaCl reading in percentage. The secondary LCD displays “100”.
% CAL
BUF
NaCl
When the reading is stable, “CFM” tag starts blinking on the LCD, asking for confirmation. Press CFM to confirm calibration.
% CAL
CFM
NaCl
BUF
CFM
The instrument displays the “Stor Good” message and returns to measurement mode.
Notes:
•
If the uncalibrated reading is too far from the expected value, calibration is not recognized. The “CFM” tag does not appear; the “~” and
“BUF” tags blink to signal wrong or contaminated calibration solution.
• The meter uses 1.90 %/°C compensation factor during calibration. If the setup item “tc” has been set to a different value, when exiting calibration mode, the displayed value on the upper LCD might be different from the nominal buffer value.
11
TEMPERATURE CALIBRATION
(for technical personnel only)
The calibration is a 2-point procedure at 0.0 and 50.0 ºC.
• Immerse the probe in a 0 ºC temperature bath.
• Press ALT+CALT keys to enter temperature calibration mode.
• The lower LCD displays “0.0 ºC”, “BUF” and “CAL” tags blinking.
• When the reading is stable, “CFM” tag starts blinking.
• Press CFM to confirm. The lower LCD displays 50.0 ºC.
• Immerse the probe in a 50 ºC temperature bath.
• When the reading is stable, “CFM” tag starts blinking.
• Press CFM to confirm. The meter returns to measurement mode.
T E M P E R A T U R E A D J U S T M E N T
The temperature reading can be manually fine-tuned following the next procedure:
Press ALT + CALT keys to enter temperature calibration mode.
ALT CALT
Press CAL to enter temperature adjustment mode. The primary and secondary LCD will display the current temperature reading.
CAL
Adjust the temperature reading on the primary
LCD using the ARROW keys. The maximum adjustment is ±1 °C around current reading.
OR
Press CFM to confirm. The meter returns to measurement mode and displays the new temperature.
CFM
Notes:
• Press ALT + CALT keys to escape without any changes.
• To enter temperature adjustment mode, the probe must be connected to the instrument.
12
C O N D U C T I V I T Y V E R S U S
TEMPERATURE CHART
The conductivity of an aqueous solution is the measure of its ability to carry an electrical current by means of ionic motion.
The conductivity invariably increases with increasing temperature.
It is affected by the type and number of ions in the solution and by the viscosity of the solution itself. Both parameters are temperature dependent. The dependency of conductivity on temperature is expressed as a relative change per
Celsius degree at a particular temperature, commonly as percent per ºC.
The following table lists the temperature dependence of the HANNA calibration buffers.
Cº Fº
2 1
2 2
2 3
2 4
1 7
1 8
1 9
2 0
1 0
1 5
1 6
0
5
2 9
3 0
3 1
2 5
2 6
2 7
2 8
6 9
7 1
7
7
7
5
3
7 7
8 8.
2.
4.
6.
8.
8
8 2
0
8
8 4
8 6
7 8.
4.
6.
2.
6
6 4
2
6 6
6 8
2.
4.
6.
3 2
4 1
5 0
6
5 9
0 8.
H
H
( µ S
7I 0 3 0
8I 0 3 0 c/ m )
H
H
( µ S
7I 0 3 1
8I 0 3 1 c/ m )
H
H
( µ S
7I 0 3 3
8I 0 3 3 c/ m )
H
H
( µ S
7I 0 3 4
8I 0 3 4 c/ m )
H
H
( µ S
7I 0 3 5
8I 0 3 5 c/ m )
H
H
( µ S
7I 0 3 9
8I 0 3 9 c/ m )
1 0 9 5 0
1 1 1 9 0
1
1 1
1
4
6
3
7 0
0
1 1 9 1 0
1 2 1 5 0
1
1 2
2
3
6
9
4 0
0
7 1 5 0
8 2 2 0
9 3 3 0
1
1 0
0
4
7
8
2 0
0
1 2 8 8 0
1 3 1 3 0
1
1 3
3
3
6
7
2 0
0
1 3 8 7 0
1 4 1 2 0
1 4 3 7 0
7 8
7 9
8 1
8 2
7 1
7 3
7 4
7 6
6 4
6 5
6 7
6 8
7 0
9 0
9 2
9 4
8 4
8 6
8 7
8 9
1 1 9 9
1 2 2 5
1 2 5 1
1 2 7 8
1 3 0 5
1 3 3 2
1 3 5 9
1 3 8 6
7 7 6
8 9 6
1 0 2 0
1 1 4 7
1 1 7 3
1 4 1 3
1 4 4 0
1 4 6 7
1 4 9 4
1 5 2 1
1 5 4 8
1 5 7 5
1 0 4 0 0 0
1 0 5 9 0 0
1
1 0
0
9
7 9 0 0
8 0 0
1 1 1 8 0 0
1 1 3 8 0 0
1
1 1
1
7
5 7 0 0
7 0 0
6 5 4 0 0
7 4 1 0 0
8 3 2 0 0
9
9 2
4
5
4
0
0 0
0
9 6 3 0 0
9 8 2 0 0
1
1 0
0
2
0 2 0 0
1 0 0
1 1 9 7 0 0
1 2 1 8 0 0
1 2 3 9 0 0
6 8 5 0 0
6 9 8 0 0
7
7
2
1 3
4
0
0 0
0
7 4 0 0 0
7 5 2 0 0
7
7 6
8
5
3
0
0 0
0
4 8 3 0 0
5 3 5 0 0
5 9 6 0 0
6
6 5
7
4
2
0
0 0
0
8 0 0 0 0
8 1 3 0 0
8
8 3
4
0
9
0
0 0
0
8 6 3 0 0
8 8 2 0 0
9 0 0 0 0
4 2 4 5
4 3 3 7
4 4 2 9
4 5 2 3
4 6 1 7
4 7 1 1
4 8 0 5
4 9 0 2
2 7 6 0
3 1 8 0
3 6 1 5
4 0 6 3
4 1 5 5
5 0 0 0
5 0 9 6
5 1 9 0
5 2 8 6
5 3 8 3
5 4 7 9
5 5 7 5
13
S E T U P
Setup mode allows viewing and modifying the instrument parameters.
To enter setup press ALT+SETUP keys when the meter is in measurement mode.
“Set” is displayed on the primary LCD.
ALT SETUP
The secondary LCD displays the code of the current setup item.
Select the desired setup item using the ARROW keys. Press CFM to confirm.
OR
CFM
Note: If ALT+SETUP keys are pressed before item confirmation, the meter exits SETUP mode and returns to measurement mode.
Once the desired setup item has been selected, its current value starts blinking
(if it is a changeable parameter).
Press the ARROW keys to change the value.
OR
Press RANGE. The other item will start blinking (e.g. month in setting up the correct date).
RANGE
Press the ARROW keys to change the value.
Press CFM to confirm.
CFM
14
Note: Press ALT+SETUP keys before confirmation to escape without changing the previously set value.
The following table lists the setup items, their valid range of values and the factory settings (default):
Item Description tc
Valid values
Temp.compensation coeff. 0.00 to 6.00 %/ºC tcE Temp.compensation mode Atc, Mtc, notc rEF Reference temperature 20.0 ºC, 25.0 ºC tdS TDS factor
CEL Cell constant (K)
AoF Auto-Off enabled
BeP Beep on key pressed
YEA Year dAt Date (DD.MM)
0.40 to 0.80
0.500 to 1.700
On, Off
On, Off
1999 to 2098
01.01 to 31.12
hou Time (hh.mm) id Meter identification code
01.01 to 23.59
0000 to 9999 vEr Firmware release x.x
Default
1.90
Atc
25.0 ºC
0.50
1.000
Off
On
1999
01.01
01.01
0000
Notes:
• Once enabled, the Auto-Off time is fixed at 5 minutes.
• Assigning an ID code is helpful in identifying a particular meter from others.
GOOD LABORATORY PRACTICE
Good Laboratory Practice (GLP) is a set of functions that allows storage and retrieval of data regarding the status of the system.
After a successful calibration, the meter automatically stores the date and time of calibration, the used calibration solution and the resulting cell constant value. All this information can be recalled by the user.
To view the last calibration data, select the desired range (EC or NaCl) and press
ALT+GLP keys.
ALT GLP
The first information appearing on the LCD is the meter “id” code.
15
By repeatedly pressing RANGE, GLP data is displayed in the following order:
Last calibration date:
RANGE
Last calibration year:
Last calibration time:
Cell constant value (K):
Offset value:
This information is displayed only if the last calibration was performed at 0.00 µS.
Calibration solution used:
BUF
If the cell constant was changed after calibration (through the “CEL” setup function), this information is not displayed.
For NaCl GLP, the last parameter is not the nominal value of the calibration solution but the actual conductivity and the temperature of the used calibration solution.
C
BUF
16
If RANGE is pressed when the last parameter is displayed, the meter returns to measurement mode.
Notes:
• To exit GLP mode at any time press ALT+GLP keys.
• If the calibration procedure was never performed, after displaying the
ID code the LCD will show the “no CAL” message blinking. Press RANGE or ALT+GLP keys to return to measurement mode.
• Last calibration data is available for EC and NaCl only. No calibration data can be recalled for TDS. If the meter is in TDS mode, by pressing
ALT+GLP keys it is possible only to view the ID code. Press ALT+GLP keys again to return to measurement mode.
• The meter has an internal lithium battery that allows to correctly update the date and time even if the power supply is disconnected.
DATA TRANSFER TO PC
Connect the meter to a PC through the RS232 output. Use HI 920010 (9 to
9-pin) connection cable.
The meter must be in measurement mode to communicate.
ALT
GLP
SETUP
ATC
TC
CAL
Functions
RANGE CFM
ATC
CFM
LOCK Start Calibration
Lock Current Range
Change V alue
Temperature
Compensation alue
SETUP
CAL
ALT Functions
TC
GLP View Calibration
Data
Start T
Enter Setup Mode
Display TC Factor
HI2300
Microprocessor Conductivity Meter
LOCK
RANGE
The meter port is optoisolated and transmits data with a Baud rate of
2400 bps.
The user can retrieve the GLP data, request meter’s parameters and the current reading (current range only) from the PC directly. It is also possible to send a command from the PC to change ranges.
To communicate with the PC use the optional HI 92000 communication software. The software is provided with an exclusive on-line guide of all the commands available and allows data printing, plotting and exporting.
17
PROBE MAINTENANCE
Rinse the probe with clean water after measurements. If more cleaning is required, remove the probe sleeve and clean the probe with a cloth or a nonabrasive detergent. Make sure to reinsert the sleeve onto the probe properly and in the right direction. After cleaning the probe, recalibrate the instrument.
The platinum rings are sustained with glass. Take great care while handling the probe.
A C C E S S O R I E S
CONDUCTIVITY BUFFER SOLUTIONS
HI 70030P 12880 µS/cm (µmho/cm), 20 mL sachets (25 pcs.)
HI 7030L 12880 µS/cm (µmho/cm), 460 mL bottle
HI 7030M 12880 µS/cm (µmho/cm), 230 mL bottle
HI 70031P 1413 µS/cm (µmho/cm), 20 mL sachets (25 pcs.)
HI 7031L 1413 µS/cm (µmho/cm), 460 mL bottle
HI 7031M 1413 µS/cm (µmho/cm), 230 mL bottle
HI 70033P 84 µS/cm (µmho/cm), 20 mL sachets (25 pcs.)
HI 7033L 84 µS/cm (µmho/cm), 460 mL bottle
HI 7033M 84 µS/cm (µmho/cm), 230 mL bottle
HI 7034L 80000 µS/cm (µmho/cm), 460 mL bottle
HI 7034M 80000 µS/cm (µmho/cm), 230 mL bottle
HI 7035L 111800 µS/cm (µmho/cm), 460 mL bottle
HI 7035M 111800 µS/cm (µmho/cm), 230 mL bottle
HI 70039P 5000 µS/cm (µmho/cm), 20 mL sachets (25 pcs.)
HI 7039L 5000 µS/cm (µmho/cm), 460 mL bottle
HI 7039M 5000 µS/cm (µmho/cm), 230 mL bottle
HI 8030L 12880 µS/cm (µmho/cm), 460 mL FDA approved bottle
HI 8031L 1413 µS/cm (µmho/cm), 460 mL FDA approved bottle
HI 8033L 84 µS/cm (µmho/cm), 460 mL FDA approved bottle
HI 8034L 80000 µS/cm (µmho/cm), 460 mL FDA approved bottle
HI 8035L 111800 µS/cm (µmho/cm), 460 mL FDA approved bottle
HI 8039L 5000 µS/cm (µmho/cm), 460 mL FDA approved bottle
HI 7037L 100% NaCl sea water standard solution, 460 mL
18
PROBE CLEANING SOLUTIONS
HI 7061M General Cleaning Solution, 230 mL bottle
HI 7061L General Cleaning Solution, 460 mL bottle
HI 8061M General Cleaning Solution, 230 mL FDA approved bottle
HI 8061L General Cleaning Solution, 460 mL FDA approved bottle
OTHER ACCESSORIES
HI 76310 Platinum 4-ring Conductivity/TDS probe with temperature sensor and 1 m (3.3') cable.
HI 710005 12VDC voltage adapter (US plug)
HI 710006 12VDC voltage adapter (European plug)
HI 920010 9 to 9-pin connection cable
HI 92000 Windows ® compatible software
RECOMMENDATIONS FOR USERS
Before using this product, make sure that it is entirely suitable for the environment in which it is used.
Operation of this instrument in residential areas could cause unacceptable interferences to radio and TV equipment, requiring the operator to follow all necessary steps to correct interferences.
The metal bonds of the probe are sensitive to electrostatic discharges. Avoid touching these metal bands at all times.
To maintain the EMC performance of this equipment, the recommended cables must be used. Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC performance.
To avoid electrical shock, do not use this instrument when voltages at the measurement surface exceed 24 VAC or 60 VDC.
To avoid damage or burns, do not perform any measurements in microwave ovens.
19
SALES AND TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66.2619.0708 • Fax 66.2619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and a complete list of Sales and
Technical offices, please see www.hannainst.com.
20
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 2 WARRANTY
- 3 PRELIMINARY EXAMINATION
- 3 GENERAL DESCRIPTION
- 4 FUNCTIONAL DESCRIPTION
- 5 SPECIFICATIONS
- 7 OPERATIONAL GUIDE
- 8 AUTORANGING
- 8 TEMPERATURE COMPENSATION
- 9 EC/TDS CALIBRATION
- 10 NaCl CALIBRATION
- 12 TEMPERATURE CALIBRATION (for technical personnel only)
- 12 TEMPERATURE ADJUSTMENT
- 13 CONDUCTIVITY VERSUS TEMPERATURE CHART
- 14 SETUP
- 15 GOOD LABORATORY PRACTICE
- 17 DATA TRANSFER TO PC
- 18 PROBE MAINTENANCE
- 18 ACCESSORIES