TEMPERATURE CALIBRATION (for technical personnel only). Hanna Instruments HI 2300
Add to My manuals20 Pages
Hanna Instruments HI 2300 is a bench microprocessor-based meter designed to measure conductivity, TDS, NaCl, and temperature. It offers a variety of features for accurate and convenient measurements, including autoranging, temperature compensation, calibration, data storage, and computer connectivity. With its high accuracy and resolution, HI 2300 is ideal for various applications in laboratories, water treatment plants, and industrial settings.
advertisement
TEMPERATURE CALIBRATION
(for technical personnel only)
The calibration is a 2-point procedure at 0.0 and 50.0 ºC.
• Immerse the probe in a 0 ºC temperature bath.
• Press ALT+CALT keys to enter temperature calibration mode.
• The lower LCD displays “0.0 ºC”, “BUF” and “CAL” tags blinking.
• When the reading is stable, “CFM” tag starts blinking.
• Press CFM to confirm. The lower LCD displays 50.0 ºC.
• Immerse the probe in a 50 ºC temperature bath.
• When the reading is stable, “CFM” tag starts blinking.
• Press CFM to confirm. The meter returns to measurement mode.
T E M P E R A T U R E A D J U S T M E N T
The temperature reading can be manually fine-tuned following the next procedure:
Press ALT + CALT keys to enter temperature calibration mode.
ALT CALT
Press CAL to enter temperature adjustment mode. The primary and secondary LCD will display the current temperature reading.
CAL
Adjust the temperature reading on the primary
LCD using the ARROW keys. The maximum adjustment is ±1 °C around current reading.
OR
Press CFM to confirm. The meter returns to measurement mode and displays the new temperature.
CFM
Notes:
• Press ALT + CALT keys to escape without any changes.
• To enter temperature adjustment mode, the probe must be connected to the instrument.
12
advertisement
Related manuals
advertisement
Table of contents
- 2 WARRANTY
- 3 PRELIMINARY EXAMINATION
- 3 GENERAL DESCRIPTION
- 4 FUNCTIONAL DESCRIPTION
- 5 SPECIFICATIONS
- 7 OPERATIONAL GUIDE
- 8 AUTORANGING
- 8 TEMPERATURE COMPENSATION
- 9 EC/TDS CALIBRATION
- 10 NaCl CALIBRATION
- 12 TEMPERATURE CALIBRATION (for technical personnel only)
- 12 TEMPERATURE ADJUSTMENT
- 13 CONDUCTIVITY VERSUS TEMPERATURE CHART
- 14 SETUP
- 15 GOOD LABORATORY PRACTICE
- 17 DATA TRANSFER TO PC
- 18 PROBE MAINTENANCE
- 18 ACCESSORIES