Agilent Technologies Electronic Counter User Manual

Below you will find brief information for HP 5328A, HP 5345A, HP 5363B. This manual covers fundamentals of time interval measurements using electronic counters. Learn about time interval averaging, error evaluations, and generating precise time intervals. Includes applications such as phase measurement and pulse train analysis.

PDF
Document
Agilent HP 5328A/5345A Electronic Counter Manual+AI Chat | Manualzz

Advertisement

Advertisement

/

Advertisement

Key features

  • Time Interval Measurement

  • Time Interval Averaging

  • Trigger Level Control

  • Input Signal Conditioning

  • Hysteresis Compensation

  • Digital Interpolation

Frequently asked questions

Time interval averaging reduces +1 or -1 count error on repetitive signals. It is useful when the factors limiting resolution are random in nature.

The slope determines if triggering occurs on rising or falling voltage. Polarity determines if triggering occurs on positive or negative voltage. The level adjusts the trigger point up or down in voltage. Attenuator reduces high amplitude input signals.

The main error sources are ±1 count error, ± trigger error, ± time base error and ± systematic error.