Mettler Toledo «Advanced Line» Transmitters Technical Specifications


Add to my manuals
20 Pages

advertisement

Mettler Toledo «Advanced Line» Transmitters Technical Specifications | Manualzz

Contents

Transmitter x100e Series for pH, Dissolved Oxygen and Conductivity

Advanced transmitters for reliable measurements and for harsh conditions

Technical Data

Short description

The cost-effective Transmitter X100 Series are suitable designed for highly reliable and accurate measurements in a wide range of industrial applications. The instruments are easy to operate and the large-size

LCD provides substantial all essential information. The measurement values are displayed in large characters and additional pictographs explain the function operation and advise any signal or functional irregularities.

Features

– Two 0/4…20 mA current outputs

– Two limit contacts

– Alarm & wash contact

– Continuous monitoring of sensor and transmitter performance

– Sensor diagnostics

– Easy operation with help of pictographs

– PID controller

– Communication with EasyClean, the METTLER TOLEDO cleaning

Drawings

Specifications transmitter pH 2100e

Specifications transmitter O

2

4100e

Specifications transmitter O

2

4100ppb

Specifications transmitter Cond 7100e

Specifications transmitter Cond Ind 7100e

Terminal assignment «Advanced Line» transmitters

General specifications «Advanced Line» transmitters

Ordering information

8

10

12

14

17

6

7

2

4

2

Drawings

Assembly

Mounting

1

2

10

9

3

4 5 6 7

144 105

27

12

11

8

1 Sealing plugs

2 Hexagon nuts

3 Pg cable glands

4 Rubber reducer

5 Pg plug

6 Enclosure screws

7 Hinge pin

8 Cable ties

9 Filler plugs

10 Gaskets

11 Washer

12 Jumper

1 Cable gland (3 pieces)

2 Breakthroughs for cable gland or conduit 1 /

2

",

Ø 21.5 mm

(2 breakthroughs).

Conduits not included!

3 Holes for post mounting

4 Holes for wall mounting

32

42

84

80

21

43

2

Drawings

Pipe mounting with ZU 0274 bracket kit

40 – 60 mm dia.

2

3

4

132 mm

1

5

Protective hood ZU 0276 for wall and pipe mounting

165 mm 132 mm

1 Protected hood

(if required)

2 Hose clamps with worm gear drive to DIN 3017

(2 pieces)

3 Postmounting plate

4 For vertical or horizontal post/pipe mounting

5 Self-tapping screws

Panel-mount kit ZU 0275 max. 25

78 27

1

5 4

2

3

1 Screws

2 Seal

3 Control panel

4 Span pieces

5 Threaded sleeves

3

4

Specifications Transmitter pH 2100e pH/mV input Input for pH, ORP electrodes or ISFET

Measurement range –1500…+1500 mV

Display range pH value –2.00…16.00

ORP –1999…+1999 mV

Glass electrode input 1)

Input resistance

Input current

Reference electrode input 1)

> 0.5 x 10

< 2 x 10

12

–12 A

Ω

Input resistance

Input current

Meas. error 1,2,3) pH value mV value

> 1 x 10 10 Ω

< 1 x 10 –10 A

< 0.02

< 1 mV

Electrode

standardization pH *

Operating modes

Zero point adjustment

Max. calibration range

-BUF Calibration with automatic buffer recognition Calimatic:

Buffer sets -01Mettler-Toledo 2.00/4.01/7.00/9.21

-02Merck/Riedel de Haen 2.00/4.00/7.00/9.00/12.00

-03Ciba (94) 2.06/4.00/7.00/10.00

-04NIST technical 1.68/4.00/7.00/10.01/12.46

-05NIST standard 1.679/4.006/6.865/9.180

-06HACH 4.00/7.00/10.18

-07WTW technical buffers 2.00/4.01/7.00/10.00

-PRD

-MAN

-DAT

±200 mV

Product Calibration

Calibration with manual entry of individual buffer values

Data entry of premeasured electrodes

Asymmetry potential: ±60 mV

Slope: 80…103 % (47.5…61 mV/pH)

Sensor standardization

ORP*

Max. calibration range

Cal timer

Sensocheck

Sensoface

ORP calibration

–700…+700 D mV

0000…9999 h automatic monitoring of glass and reference electrode (can be disabled) provides information on the electrode condition.

Evaluation of zero/slope, response, calibration interval, Sensocheck

Specifications

Temperature input *

Measurement range

Adjustment range

Resolution

Meas. error 1,2,3)

Temp. compensation of process medium

Power output

Transmitter pH 2100e

Pt100/Pt1000/NTC 30 kΩ / NTC 8,55 kΩ

2-wire connection, adjustable

Pt100/Pt1000:

NTC 30 kΩ

NTC 8.55 kΩ

–20.0…+ 200.0 °C/–4…+ 392 °F

– 20.0…+ 150.0 °C/–4…+ 302 °F

–10.0…+ 130.0 °C/+14…+ 266 °F

10 K

0.1 °C/1 °F

< 0.5 K (< 1 K for Pt100; <1K for NTC >100 °C)

Linear –19.99…+19.99 %/K

(reference temp. 25 °C) for operating an ISFET adapter

+ 3 V/0.5 mA

– 3 V/0.5 mA

* User-defined

1) To IEC 746 Part 1, at nominal operating conditions

2)

3)

± 1 count

Plus sensor error

5

6

Specifications Transmitter O

2

4100e

Dissolved oxygen input

Sensor type A:

Sensor type B:

Measuring current

Resolution

InPro6000…6800

InPro6900

-2…1800 nA

0.05 nA (with Vpol ≤ 800 mV and Vref ≤ 200 mV)

Saturation (–10…80 °C) 0…500 %

Meas. error 1,2,3) 0.5 % of meas. val. +0.5 %

Concentration (–10…80 °C) 0.00…50.00 mg/l

0.00…50.00 ppm

Meas. error 1,2,3) 0.5 % of meas. val. + 0.05 mg/l

Adm. guard current

Polarization voltage *

Process pressure* or 0.05 ppm

20 µA

0…1000 mV

0.000…9.999 bar

Salt correction *

(…999.9 kPa/…145.0 psi)

00.00…45.00 g/kg

Sensor standardization

Operating modes *

Calibration range

Sensor type A

Calibration range

Sensor type B

Calibration timer *

Pressure correction *

DO saturation (automatic)

DO concentration (automatic)

Product calibration

Zero point calibration

Zero point

Slope

± 2 nA

25…130 nA

(at 25 °C, 1013 mbars)

± 2 nA Zero point

Slope 200…550 nA

(at 25 °C, 1013 mbars)

0000…9999 h

0.000…9.999 bars/999.9 kPa/145.0 psi

Sensocheck

Sensoface

Monitoring for short circuits/open circuits (can be disabled)

Provides information on the sensor condition

Evaluation of zero/slope, response, calibration interval, Sensocheck

Temperature input *

Measurement range

Adjustment range

Resolution

Meas. error 1,2,3)

NTC 22 kΩ / NTC 30 kΩ*

2-wire connection, adjustable

– 20.0…+150.0 °C/– 4…+ 302 °F

10 K

0.1 °C/1 °F

< 0.5 K (< 1 K at >100 °C)

* User-defined

1) To IEC 746 Part 1, at nominal operating conditions

2)

3)

± 1 count

Plus sensor error

Specifications Transmitter O

2

4100ppb

Dissolved oxygen input

Sensor type A:

Sensor type B:

Measuring current

Resolution

InPro6000…6800

InPro6900

-2…600 nA

0.01 nA (with Vpol ≤ 500 mV and Vref ≤ 200 mV)

Saturation (–10…80 °C) 0.0.…120.0 %

Meas. error 1,2,3) 0.5 % of meas. val. +0.1 %

Concentration (–10…80 °C) 0000…9999 µg/l

0000…9999 ppb

0.0000…9.999 mg/l

0.0000…9.999 ppm

Meas. error 1,2,3) 0.5 % of meas. val. +0.005 mg/l or 0.005 ppm

Adm. guard current

Polarization voltage*

Process pressure*

Salt correction*

20 µA

0…1000 mV

0.000…9.999 bars

(to 999.9 kPa/…145.0 psi)

00.00…45.00 g/kg

Sensor standardization

Operating modes *

Calibration range

Sensor type A

Calibration range

Sensor type B

Calibration timer *

Pressure correction *

Sensocheck

Sensoface

DO saturation (automatic)

DO concentration (automatic)

Product calibration

Zero point calibration

Zero point

Slope

± 2 nA

25…130 nA

(at 25 °C, 1013 mbars)

± 2 nA Zero point

Slope 200…550 nA

(at 25 °C, 1013 mbars)

0000…9999 h

0.000…9.999 bars/999.9 kPa/145.0 psi

Monitoring for short circuits/open circuits (can be disabled)

Provides information on the sensor condition

Evaluation of zero/slope, response, calibration interval, Sensocheck

Temperature input *

Measurement range

Adjustment range

Resolution

Meas. error 1,2,3)

NTC 22 kΩ/ NTC 30 kΩ*

2-wire connection, adjustable

–20.0…+150.0 °C/–4…+302 °F

10 K

0.1 °C/1 °F

< 0.5 K (< 1 K at >100 °C)

* User-defined

1) To IEC 746 Part 1, at nominal operating conditions

2)

3)

± 1 count

Plus sensor error

7

8

Specifications

Conductivity input

Working range

Effective ranges

Measurement error 1,2,3)

Conc. measurements

Sensor standardization

Permissible cell constant

Sensocheck

Sensoface

Sensor monitor

USP-Function

Transmitter Cond 7100e

Input for 2-e or 4-e conductivity sensors

4-electrode sensor: 0.2 µS *c …1000 mS * c (c = cell constant)

2-electrode sensor: 0.2 µS *c …200 mS * c

(the actual range is very much depending on the sensor used, display limited to 3500 mS)

Conductivity 0.000…9.999 µS/cm

00.00…99.99 µS/cm

000.0…999.9 µS/cm

0000…9999 µS/cm

0.000…9.999 mS/cm

00.00…99.99 mS/cm

000.0…999.9 mS/cm

0.000…9.999 S/m

00.00…99.99 S/m

Resistivity

Concentration

00.00…99.99 MΩcm

0.00…9.99 % by wt.

Salinity 0.0…45.0 ‰ (0…35 °C)

< 1% of measured value +0.4 µS * c

-01- NaCl

-02- HCl

0.00…9.99 % by wt.

(0 …60°C)

0.00…9.99 % by wt.

(–20…50 °C)

-03- NaOH 0.00…9.99 % by wt.

(0…100 °C)

-04- H

2

SO

-05- HNO

3

4

0.00…9.99 % by wt.

(–17…110 °C)

0.00…9.99 % by wt.

(–17…50 °C)

Input of cell constant with simultaneous display of conductivity value and temperature

Input of conductivity value with simultaneous display of cell constant and temperature

Product calibration

Temperature probe adjustment

00.0050…19.9999 cm –1

Monitoring of sensor polarization and cable capacitance (can be disabled)

Provides information on the sensor condition (Sensocheck)

Display of direct measurement values for validation purpose (resistance/temperature)

Monitoring of conductivity of water for pharmaceutical applications to USP (USP <645>) with adjustable limit values (10…100 % of USP value)

Specifications Transmitter Cond 7100e

Temperature input

Measuring range

*) Pt100/Pt1000/NTC 30 kΩ / NTC 8.55 kΩ

2-wire connection, adjustable

Pt100/Pt1000

NTC 30 kΩ

NTC 8.55 kΩ

0.1 °C / 1 °F

–20…+200 °C / –4…+392 °F

–20…+150 °C / -4…+302 °F

–10…+130 °C / +14…+266 °F

Resolution

Error 1,2,3) 0.5 K (< 1 K with Pt100; < 1 K with NTC > 100 °C)

Temperature compensation (OFF) not compensated

(ref. temp. 25 °C) (Lin) linear, 0.00…19.99 %/K, –20…130 °C

(NLF) natural waters to EN 27888, 0…36 °C

(nACL) ultrapure water with NaCl traces, 0…120 °C

(HCL) ultrapure water with HCl traces,

(nH3) ultrapure water with NH

3 traces,

0…120 °C

0…120 °C

* User-defined

1) To IEC 746 Part 1, at nominal operating conditions

2)

3)

± 1 count

Plus sensor error

9

10

Specifications

Conductivity input

Working range

Effective ranges

Measurement error 1,2,3)

Conc. measurements

Sensor standardization

Transmitter Cond Ind 7100e

Input for inductive sensors

Conductivity 0.000…1999 mS/cm

Concentration

Salinity

Conductivity

0.00…100 % by wt.

0.0…45.0 ‰ (0…35 °C)

0.000…9.999 mS/cm

00.00…99.99 mS/cm

000.0…999.9 mS/cm

0000…1999 mS/cm

0.000…9.999 S/m

00.00…99.99 S/m

Concentration

Salinity

0.00…99.99 %

0.0…45.0 ‰ (0…35 °C)

< 1% of measured value +0.005 mS

-01- NaCl

-02- HCl

0…26 % (0…60 °C)

0…18 % (–20…50 °C)

-03- NaOH 0…14 % (0…100 °C)

-04- H

2

SO

4

-05- HNO

3

-06- H

2

SO

-07- HCl

4

0…30 % (–17…110 °C)

0…30 % (–20…50 °C)

92…99 % (–17…115 °C)

22…39 % (–20…50 °C)

-08- HNO

3

-09- H

2

SO

4

35…96 % (–20…50 °C)

32…84 % (–17…115 °C)

-10- NaOH 18…50 % (0…100 °C)

Input of cell factor with simultaneous display of conductivity value and temperature

Input of conductivity value with simultaneous display of cell factor and temperature

Product calibration

Zero point calibration

Temperature probe adjustment

00.100…19.999

01.00…199.99

Permissible cell factor

Permissible transfer ratio

Permissible zero point deviation

Sensocheck

Sensoface

Sensor monitor

± 0.5 mS/cm

Monitoring of sender coil and leads for short circuiting, and of the receiver coil for disruption (can be disabled)

Indicates sensor status (zero point, sensocheck)

Display of direct measurement values for validation purpose (resistance /temperature)

Specifications Transmitter Cond Ind 7100e

Temperature input

Measuring range

* Pt100/Pt1000/NTC 100 kΩ

2-wire connection, adjustable

Pt100/Pt1000 –20…+200 °C / –4… +392 °F

NTC –20…+130 °C / –4… +266 °F

0.1 °C / 1 °F Resolution

Error 1,2,3) 0.5 K (< 1 K with Pt100; < 1 K with NTC > 100 °C)

Temperature compensation (OFF) not compensated

(ref. temp. 25 °C) (LIN) linear, 0.00…19.99 %/K

(NLF) natural waters to EN 27888 (0…35 °C)

* User-defined

1)

2)

To IEC 746 Part 1, at nominal operating conditions

± 1 count

3) Plus sensor error

11

12

Terminal assignment

Transmitter pH 2100 e

Transmitter O

2

4100 e

Transmitter O

2

4100 ppb

«Advanced Line» transmitters

Terminal assignment

Transmitter Cond 7100e

Transmitter Cond Ind 7100e

«Advanced Line» transmitters

13

14

General specifications

HOLD input

Function

Switching voltage

CONTROL input pH/O

2

Function

)

Switching voltage

CONTROL input (Cond)

Function

Switching voltage

Output 1

Process variable *

Current characteristics*

Overrange *

Output filter *

Meas. error

1)

Start/end of scale

Adm. span

Output 2

Process variable

Overrange *

Output filter *

Meas. error 1)

Start/end of scale *

Adm. span

Alarm contact

Contact ratings

Contact response

Alarm delay

«Advanced Line» transmitters

Galv. separated (OPTO coupler)

Switches device to HOLD state

0…2 V (AC/DC) hold inactive

10…30 V (AC/DC) hold active

Galv. separated (OPTO coupler)

Control input for automatic cleaning/calibration system

0…2 V (AC/DC) inactive

10…30 V (AC/DC) active

Galv. separated (OPTO coupler)

Switches between two parameter sets

0…2 V (AC/DC) set #1 active

10…30 V (AC/DC) set #2 active

0/4…20 mA, max. 10 V, floating

(galv. connected to output 2) pH 2100 e

O

2

O

2

4100e

4100 ppb

Cond 7100e pH/mV

%, mg/l

%, mg/l conductivity, resistivity, concentration, salinity

Cond Ind 7100e conductivity, concentration, salinity linear or logarithmic (depends on transmitter)

22 mA in the case of error message

Low-pass, filter time constant 0…120 s

< 0.3 % of current value +0.05 mA

As desired within measuring range pH 2100e

0

2

0

2

4100e

4100 ppb

Cond 7100e

2.00…18.00/200…3000 mV

5…500 %, 0.5…50 mg/l

2…200 %, 0.2…10 mg/l

LIN 5 % of selected measuring range

LOG 1 decade

Cond Ind 7100e LIN 5 % of selected measuring range

LOG 1 decade

0/4…20 mA, max. 10 V, floating

(galv. connected to output 1)

Temperature

22 mA in the case of temp error messages

Low-pass, filter time constant 0…120 s

< 0.3 % of current value + 0.05 mA pH, Cond: –20…+200 °C/–4…+392 °F, 0

2 pH, Cond: 20…320 K, 0

2

: 20…170 K

: –20…+150 °C/–4...+302 °F

Relay contact, floating

AC < 250 V/< 3 A/< 750 VA

DC < 30 V/< 3 A/< 90 W

N/C (fail-safe type)

0000…0600 s

General specifications «Advanced Line» transmitters

Limit values

Contact ratings

Contact response *

Delay *

Switching points *

Hysteresis *

PID process controller

Setpoint specification *

Neutral zone *

P-action *

I-action component *

D-action component *

Output via relay contacts R1, R2

Contacts R1, R2 floating, but inter-connected

AC < 250 V/< 3 A/< 750 VA

DC < 30 V/< 3 A/< 90 W

N/C or N/O

0000…9999 s

As desired within measuring range pH 2100e 0…5.00 pH/0…500 mV

0

2

0

2

4100e

4100ppb

Cond 7100e

0…50 %/0…5.00 mg/l (ppm)

0…50 %/0…5.00 mg/l (ppm)

0…50 % of measuring range

Cond Ind 7100e 0…50 % of measuring range

Output via relay contacts R1, R2 (see limit values) pH 2100e –02.00…16.00/– 1500…+1500 mV

0

2

0

2

4100e

4100ppb

Cond 7100e

0…500 % / 0…50 mg/l

0…120% / 0…9.999 mg/l within selected measuring range

Cond Ind 7100e within selected measuring range pH 2100e

0

2

0

2

4100e

4100ppb

Cond 7100e

0…5.00 pH / 0…+500 mV

0…50 % / 0…5 mg/l

0…50 % / 0…5 mg/l (ppm) within selected measuring range

Cond Ind 7100e max. 50 % of selected measuring range

Controller gain K

Reset time Tr:

R

: 0010…9999 %

0000…9999 s

(0000 s = no integral action)

Derivative-action time Td: 0000…9999 s

(0000 s = no derivative action)

Pulse length controller or pulse frequency controller

0001…0600 s, min. ON time 0.5 s (pulse length controller)

0001…0180 min –1 (pulse frequency controller)

Controller type*

Pulse period *

Max. pulse frequency *

Cleaning function/

2 nd parameter set

Contact ratings

Contact response *

Rinsing interval *

Cleaning time *

Calibration interval *

Cleaning interval*

Relay contact, floating for controlling a simple rinsing system or an automatic cleaning system or to show that 2 nd parameter set is active

AC < 250 V/< 3 A/< 750 VA

DC < 30 V/< 3 A/< 90 W

N/C or N/O (cleaning function)

N/O (2 nd parameter set)

000.0…999.9 h

(000.0 h = cleaning function switched off)

0000…1999 s

000.0…999.9 h

000.0…999.9 h

* User-defined

1)

2)

To IEC 746 Part 1, at nominal operating conditions

± 1 count

3) Plus sensor error

15

16

General specifications «Advanced Line» transmitters

Display

Main display

Secondary display

Sensoface

Mode indicators

Alarm indication

Keypad

Service functions

Current source

Manual controller

Device self-test

Display test

Last Error

Sensor monitor

Relay test

Parameter sets*

Data retention

EMC

Lightning protection

LC display, 7-segment with icons

Character height 17 mm, unit symbols 10 mm

Character height 10 mm, unit symbols 7 mm

3 status indicators (friendly, neutral, sad smiley)

5 status bars: «meas», «cal», «alarm», «cleaning», «config»

18 further icons for configuration and messages

Red LED in case of alarm or HOLD, user defined

5 keys: [cal] [conf] [ ] [ ] [enter]

Current specifiable for output 1 and 2 (00.00…22.00mA)

Controller output entered directly (start of control process)

Automatic memory test (RAM, FLASH, EEPROM)

Display of all segments

Display of last error occurred

Display of direct, uncorrected sensor signal (electrode/sensor)

Manual control of the four switching contacts

Two selectable parameter sets for different applications

Parameters and calibration data > 10 years (EEPROM)

EN 61326

EN 61326/A1

EN 61000-4-5, Installation Class 2

Protection against electrical shock

FM/CSA

Power supply

Protective separation of all extra-low-voltage circuits against mains as per EN 61010

NI, Class 1, Div 2, Group A, B, C, D, T4

24 (–15%)…230 (+10%) V AC/DC; approx. 5 VA, 2.5 W

AC: 45…65 Hz

Overvoltage category II, Class II

Nominal operating conditions

Ambient temperature – 20…+55 °C / –4...+131 °F

Transport/Storage temp

Relative humidity

– 20…+70 °C / -4...+158 °F

10…95 % non condensing

Power supply

Frequency for AC

24 (–15%)…230 (+ 10 %) V AC/DC

45…65 Hz

General specifications and ordering information «Advanced Line» transmitters

Enclosure

Color

Assembly

Dimensions

Ingress protection

Cable glands

Weight molded enclosure made of PBT (polybutylene terephtalate)

Bluish gray RAL 7031

• Wall mounting

• Pipe mounting: Ø 40…60 mm, Ø 30…45 mm

• Panel mounting, cutout to DIN 43 700, sealed against panel

H x W x L: 144 x 144 x 105 mm (5.67 x 5.67 x 4.13")

IP 65/NEMA 4X

3 breakthroughs for cable glands M20 x 1.5

2 breakthroughs for NPT approx. 1 kg

* User-defined

1) To IEC 746 Part 1, at nominal operating conditions

2)

3)

± 1 count

Plus sensor error

1 /

2

" or Rigid Metallic Conduit

Ordering information

Description

4-wire instruments (20…253 V AC/DC)

Transmitter pH 2100e

Transmitter O

Transmitter O

2

4100e

2

4100ppb

Transmitter Cond 7100e

Transmitter Cond Ind 7100e

Installation accessories

Bracket kit

Panel-mount kit

Protective hood

Designation pH 2100e

O

2

O

2

4100e

4100ppb

Cond 7100e

Cond Ind 7100e

ZU 0274

ZU 0275

ZU 0276

Order number

52 121 102

52 121 103

52 121 104

52 121 126

52 121 127

52 120 741

52 120 740

52 120 739

17

18

Notes

Notes

19

METTLER TOLEDO Market Organizations

Sales and Service:

Australia

Mettler-Toledo Ltd.

220 Turner Street

Port Melbourne

AUS-3207 Melbourne/VIC

Phone +61 1300 659 761

Fax +61 3 9645 3935 e-mail [email protected]

Austria

Mettler-Toledo Ges.m.b.H.

Südrandstraße 17

A-1230 Wien

Phone +43 1 604 19 80

Fax +43 1 604 28 80 e-mail [email protected]

Brazil

Mettler-Toledo Ind. e Com. Ltda.

Avenida Tamboré, 418

Tamboré

BR-06460-000 Barueri/SP

Tel.

Fax

+55 11 4166 7400

+55 11 4166 7401 e-mail [email protected]

[email protected]

China

Mettler-Toledo Instruments

(Shanghai) Co. Ltd.

589 Gui Ping Road

Cao He Jing

CN-200233 Shanghai

Phone +86 21 64 85 04 35

Fax +86 21 64 85 33 51 e-mail [email protected]

Croatia

Mettler-Toledo d.o.o.

Mandlova 3

HR-10000 Zagreb

Phone +385 1 292 06 33

Fax +385 1 295 81 40 e-mail [email protected]

Czech Republic

Mettler-Toledo s.r.o.

Trebohosticka 2283/2

CZ-100 00 Praha 10

Phone +420 2 72 123 150

Fax +420 2 72 123 170 e-mail [email protected]

Denmark

Mettler-Toledo A/S

Naverland 8

DK-2600 Glostrup

Phone +45 43 27 08 00

Fax +45 43 27 08 28 e-mail [email protected]

France

Mettler-Toledo

Analyse Industrielle S.A.S.

30, Boulevard de Douaumont

F-75017 Paris

Phone +33 1 47 37 06 00

Fax +33 1 47 37 46 26 e-mail [email protected]

Germany

Mettler-Toledo GmbH

Prozeßanalytik

Ockerweg 3

D-35396 Gießen

Phone +49 641 507 333

Fax +49 641 507 397 e-mail [email protected]

Great Britain

Mettler-Toledo LTD

64 Boston Road, Beaumont Leys

GB-Leicester LE4 1AW

Phone +44 116 235 7070

Fax +44 116 236 5500 e-mail [email protected]

Hungary

Mettler-Toledo Kereskedelmi KFT

Teve u. 41

HU-1139 Budapest

Phone +36 1 288 40 40

Fax +36 1 288 40 50 e-mail [email protected]

India

Mettler-Toledo India Private Limited

Amar Hill, Saki Vihar Road

Powai

IN-400 072 Mumbai

Phone +91 22 2857 0808

Fax +91 22 2857 5071 e-mail [email protected]

Italy

Mettler-Toledo S.p.A.

Via Vialba 42

I-20026 Novate Milanese

Phone +39 02 333 321

Fax +39 02 356 2973 e-mail [email protected]

Japan

Mettler-Toledo K.K.

Process Division

6F Ikenohata Nisshoku Bldg.

2-9-7, Ikenohata

Taito-ku

JP-110-0008 Tokyo

Phone +81 3 5815 5606

Fax +81 3 5815 5626 e-mail [email protected]

Malaysia

Mettler-Toledo (M) Sdn Bhd

Bangunan Electroscon Holding, U 1-01

Lot 8 Jalan Astaka U8/84

Seksyen U8, Bukit Jelutong

MY-40150 Shah Alam Selangor

Phone +60 3 78 44 58 88

Fax +60 3 78 45 87 73 e-mail

[email protected]

Mexico

Mettler-Toledo S.A. de C.V.

Ejercito Nacional #340

Col. Chapultepec Morales

Del. Miguel Hidalgo

MX-11570 México D.F.

Phone +52 55 1946 0900 e-mail [email protected]

Poland

Mettler-Toledo (Poland) Sp.z.o.o.

ul. Poleczki 21

PL-02-822 Warszawa

Phone +48 22 545 06 80

Fax +48 22 545 06 88 e-mail [email protected]

Russia

Mettler-Toledo Vostok ZAO

Sretenskij Bulvar 6/1

Office 6

RU-101000 Moscow

Phone +7 495 621 56 66

Fax +7 495 621 63 53 e-mail [email protected]

Singapore

Mettler-Toledo (S) Pte. Ltd.

Block 28

Ayer Rajah Crescent #05-01

SG-139959 Singapore

Phone +65 6890 00 11

Fax +65 6890 00 12

+65 6890 00 13 e-mail [email protected]

Slovakia

Mettler-Toledo s.r.o.

Hattalova 12/A

SK-83103 Bratislava

Phone +421 2 4444 12 20-2

Fax +421 2 4444 12 23 e-mail [email protected]

Slovenia

Mettler-Toledo d.o.o.

Pot heroja Trtnika 26

SI-1261 Ljubljana-Dobrunje

Phone +386 1 530 80 50

Fax +386 1 562 17 89 e-mail [email protected]

South Korea

Mettler-Toledo (Korea) Ltd.

Yeil Building 1 & 2 F

124-5, YangJe-Dong

SeCho-Ku

KR-137-130 Seoul

Phone +82 2 3498 3500

Fax +82 2 3498 3555 e-mail [email protected]

Spain

Mettler-Toledo S.A.E.

C/Miguel Hernández, 69-71

ES-08908 L’Hospitalet de Llobregat

(Barcelona)

Phone +34 902 32 00 23

Fax +34 902 32 00 24 e-mail [email protected]

Sweden

Mettler-Toledo AB

Virkesvägen 10

Box 92161

SE-12008 Stockholm

Phone +46 8 702 50 00

Fax +46 8 642 45 62 e-mail [email protected]

Switzerland

Mettler-Toledo (Schweiz) GmbH

Im Langacher

Postfach

CH-8606 Greifensee

Phone +41 44 944 45 45

Fax +41 44 944 45 10 e-mail [email protected]

Thailand

Mettler-Toledo (Thailand) Ltd.

272 Soi Soonvijai 4

Rama 9 Rd., Bangkapi

Huay Kwang

TH-10320 Bangkok

Phone +66 2 723 03 00

Fax +66 2 719 64 79 e-mail

[email protected]

USA/Canada

Mettler-Toledo Ingold, Inc.

36 Middlesex Turnpike

Bedford, MA 01730, USA

Phone +1 781 301 8800

Freephone +1 800 352 8763

Fax e-mail

+1 781 271 0681 [email protected]

ISO

9001 certified

ISO

14001 certified

Management System certified according to

ISO 9001 / ISO 14001

Subject to technical changes.

© Mettler-Toledo AG, Process Analytics

08/13 Printed in Switzerland. 52 121 132

Mettler-Toledo AG, Process Analytics

Im Hackacker 15, CH–8902 Urdorf

Tel. + 41 44 729 62 11, Fax +41 44 729 66 36 www.mt.com

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

advertisement