Mettler Toledo «Advanced Line» 2-Wire and FF Transmitters for pH, O2 and Conductivity Technical Specifications
Add to My manuals20 Pages
advertisement
![Mettler Toledo «Advanced Line» 2-Wire and FF Transmitters for pH, O2 and Conductivity Technical Specifications | Manualzz Mettler Toledo «Advanced Line» 2-Wire and FF Transmitters for pH, O2 and Conductivity Technical Specifications | Manualzz](http://s3.manualzz.com/store/data/053267240_1-4627a69a6b35ac6c4ec144cc098fc9c0-360x466.png)
Contents
«Advanced Line» 2-wire Transmitter, for pH, O
2
and Conductivity
Transmitters for reliable measurements and for harsh conditions in Ex and non-Ex versions with HART
®
, Profibus
® and FOUNDATION
™
Fieldbus interface.
Technical Data
Short description
The cost-effective 2-wire transmitters are suitable designed for highly reliable and accurate measurements in a wide range of industrial applications. The instruments are easy to operate and the large-size LCD provides substantial all essential information. The measurement values are displayed in large characters and additional pictographs explain the function operation and advise any signal or functional irregularities.
Features
– ATEX/FM approval for Ex instruments
– Product calibration
– Continous monitoring of sensor and transmitter performance
– Sensor diagnostics
– Easy to operate with help of pictographs
– HART ® , Profibus ® PA and FOUNDATION ™ Fieldbus communication
– FDA 21 CFR Part 11 conformity
HART
®
FIELD COMMUNICATION PROTOCOL
Drawings
Specifications pH 2100 e / 2(X)H, Profibus pH 2100 PA, Fieldbus pH 2100 e FF
Specifications O
2
4100 e / 2(X)H, Profibus O
2
4100 PA, Fieldbus O
2
4100 e FF
Specifications Cond 7100 e / 2(X)H, Profibus Cond 7100 PA, Fieldbus Cond 7100 e FF
Specifications Cond Ind 7100 e / 2(X)H, Profibus Cond Ind 7100 PA, Fieldbus Cond Ind 7100 e FF
Terminal assignment «Advanced Line» transmitters
General specifications «Advanced Line» transmitters
Ordering information
10
12
16
18
6
8
2
4
2
Drawings
Assembly
12
11
10
9
8
7 6 5
4
1
2
3
1 Sealing plugs (1 piece)
2 Hexagon nuts (2 pieces)
3 Pg cable glands (3 pieces)
4 Rubber reducer (1 piece)
5 Pg plug (1 piece)
6 Enclosure screws (4 pieces)
7 Hinge pin (1 piece)
8 Cable ties (3 pieces)
9 Filler plugs (3 pieces)
10 Gaskets (3 pieces)
11 Washer (5 pieces)
12 Jumper (2 pieces)
Mounting
32
42
84
80
144 105
27
1 Cable gland (3 pieces)
2 Breakthroughs for cable gland or conduit
1
/
2
",
Ø 21.5 mm
(2 breakthroughs).
Conduits not included!
3 Holes for post mounting
4 Holes for wall mounting
All dimensions in mm
21
43
2
Drawings
Pipe mounting with ZU 0274 bracket kit
40 – 60 mm dia.
2
3
4
132 mm
1
5
Protective hood ZU 0276 for wall and pipe mounting
165 mm 132 mm
1 Protective hood
(if required)
2 Hose clamps with worm gear drive to DIN 3017
(2 pieces)
3 Pipe mount plate
(1 piece)
4 For vertical or horizontal post/pipe mounting
5 Self-tapping screws
(4 pieces)
Panel-mount kit ZU 0275
max. 25
78 27
1
5
4
2
3
1 Screws (4 pieces)
2 Seal (1 piece)
3 Control panel (4 pieces)
4 Span pieces
5 Threaded sleeves (4 pieces)
All dimensions in mm
3
4
Specifications
Transmitters pH/mV input
Sensor standardization pH
*)
Operating modes
«Advanced Line» transmitters for pH measurement
2-wire HART pH 2100 e / 2(X)H
FOUNDATION™ Fieldbus pH 2100 e FF
Profibus PA pH 2100 PA
Input pH or ORP electrodes or ISFET
4)
Measurement range –1500 ... +1500 mV
–1500 ... +1000 mV for FF version
Display range pH value –2.00 ... 16.00
ORP: –1999 ... +1999 mV
ORP: –1500 ... +1000 mV for FF version
Glass electrode input
1)
Input resistance > 0.5 x 10
12
< 2 x 10
–12
A
Ohms
Input current
Reference electrode input
1)
Input resistance > 1 x 10
10
Ohms
< 1 x 10
–10
A Input current error
Measurement error
1,2,3) pH value/mV value: < 0,02 / < 1 mV pH calibration
-BUF Calibration with Calimatic automatic buffer recognition:
Buffer sets -01- METTLER TOLEDO 2.00 / 4.01 / 7.00 / 9.21
-02-
-03-
-04-
Merck / Riedel de Haen 2.00 / 4.00 / 7.00 / 9.00 / 12.00
Ciba (94) 2.06 / 4.00 / 7.00 / 10.0
NIST technical 1.68 / 4.00 / 7.00 / 10.01 / 12.46
Zero offset
Max.calibration range
Sensor standardization ORP
*)
Calibration timer
Sensocheck
Sensoface
Sensor monitor
-MAN
-DAT
-PRD
± 200 mV
-07- WTW techn. buffer 2.00 / 4.01 / 7.00 / 10.00
Manual calibration with input of individual buffer values
Data entry of premeasured electrodes
Product calibration
Asymmetry potential: ± 60 mV
Slope: 80 ... 103 % (47.5 ... 61 mV/pH)
ORP calibration max. calibration range –700 ... +700 mV
0000 ... 9999 h
Automatic monitoring of glass and reference electrode (can be disabled)
Provides information on the electrode status
Evaluation of zero / slope, response, calibration interval, Sensocheck
Direct display of measured values from sensor for validation resistance / temperature
Specifications «Advanced Line» transmitters for pH measurement
Temperature input
*)
Temp. compensation of process medium
Power output
Pt 100 / Pt 1000 / NTC 30 kOhm / NTC 8.55 kOhm
2-wire connection, adjustable
Measurement range
Pt 100/Pt 1000
NTC 30 kOhm
NTC 8.55 kOhm
– 20.0 ... + 200.0 °C (– 4 ... + 392 °F)
– 20.0 ... + 150.0 °C (– 4 ... + 302 °F)
– 20.0 ... + 130.0 °C (– 4 ... + 266 °F)
10 K Adjustment range
Resolution
Measurement error
1,2,3)
0.1 °C / 1 °F
< 0.5 K (< 1 K with Pt100;
< 1K with NTC >100 °C)
Linear – 19.99 ... + 19.99 % / K (reference temperature 25 °C)
For operating an ISFET adapter
4)
+ 3 V / 0.5 mA
– 3 V / 0.5 mA
*) User-defined
1) According to IEC 746 part 1, at nominal operating conditions
2) ± 1 count
3) Plus sensor error
4) Not valid for pH 2100 PA
5
6
Specifications «Advanced Line» transmitters for O
2 measurement
Transmitters
O
2 input
2-wire HART
FOUNDATION™ Fieldbus
O
2
4100 e / 2(X)H
O
2
4100 e FF
Profibus PA O
2
4100 PA
Sensor Typ A
Sensor Typ B
InPro 6800
InPro 6900
O
2
4100e /2(X)H
Measuring current
Saturation (–10 ... 80 °C)
Meas. error
1,2,3)
0 ... 1200 nA, resolution 20 pA
0.0 ... 199.9 % / 200 ... 500 %
0.5 % meas. val. +0.05 nA, TC: 0.005 nA/K
Concentration (–10 ... 80 °C) 0.00 ... 50.00 mg/l 0.00 ... 50.00 ppm
0000 ... 9999 µg/l 0000 ... 9999 ppb
Volume concentration in gas
(–10 ... 80 °C) 0500 ... 9999 ppm
Adm. guard current
0 % ... 120 %
≤
20 µA
O
2
4100 PA, O
2
4100 e FF
Ranges
Measuring current
Saturation
*)
Meas. error
1,2,3)
Concentration
Meas. error
1,2,3)
Measuring range 1 (low level) Measuring range 2 (high level)
–2 … 600 nA, resolution 10 pA –2 … 1800 nA, resolution 30 pA
0.0 ... 120.0 % 0 ... 500 %
0.5 % meas. val. +0.1 % sat.
0.5 % meas. val. +0.5 % saturation
0000 ... 9999 µg/l
0000 ... 9999 ppb
0.000 ... 9.999 ppm
0.000 ... 9.999 mg/l
0.0 ... 50.0 mg/l
0.0 ... 50.0 ppm
0.5 % meas. val. + 5 µg/l or 5 ppb
0.05 % meas.val. + 0.05 mg/l or 0.05 ppm
Volume concentration in gas
4)
(–10...80 °C)
Meas. error
1, 2, 3)
0000 ... 9999 ppm
0.00 % ... 120.0 %
(0.00 % ... 29.99 %
30.0 % ... 120.0 %)
0.5 % meas.val.
+ 0.02 % or 200 ppm
0000 ... 9999 ppm
0.00 % ... 120.0 %
(0.00 % ... 29.99 %
30.0 % ... 120.0 %)
0.5 % meas.val.+0.1
or 1000 ppm
Sensor standardization
Polarization voltage
*)
Process pressure
*)
Salinity correction
*)
Operating modes
Calibration range
Sensor Type A
Calibration range
Sensor Type B
Calibration timer
*)
Pressure correction
*)
(0) 400 ... 1000 mV / 10 mV steps
0.000 ... 9.999 bars (... 999.9 kPa / ... 145.0 psi)
00.00 ... 45.00 g/kg
O
2 saturation (automatic), O
2 concentration
(automatic), Volume concentration in gas
4)
,
Product calibration, Zero calibration
Zero point ±2 nA
Slope 25 ... 130 nA (at 25 °C, 1013 mbar)
Zero point ±2 nA
Slope 200 ... 550 nA (at 25 °C, 1013 mbar)
0000 ... 9999 h
0.000 ... 9.999 bar (… 999.9 kPa / …145.0 psi)
Specifications
Sensocheck
Sensoface
Temperature input
*)
«Advanced Line» transmitters for O
2 measurement
Monitoring for short cirucits / open circuits (can be disabled)
Provides information on the condition of the sensor, evaluation of zero point/slope, response time, calibration interval, Sensocheck
NTC 22 kOhm / NTC 30 kOhm
*)
2-wire connection, adjustable
Measurement range – 20.0 to +150.0 °C (– 4 to + 302 °F)
Adjustment range
Resolution
Measurement error
1, 2, 3)
10 K
0.1 °C / 1 °F
< 0.5 K (<1 K at 100 °C)
*) User-defined
1) According to IEC 746 part 1, at nominal operating conditions
2) ±1 count
3) Plus sensor error
4) Not valid for O
2
4100 PA
7
8
Specifications «Advanced Line» transmitters for conductivity measurement
Transmitters
Conductivity input
Working ranges
Effective ranges
Concentration
*)
4)
Sensor standardization
Permissible cell constant
USP function
4)
Sensocheck
Sensoface
Sensor monitor
Temperature input
*)
2-wire HART
FOUNDATION™ Fieldbus
Profibus PA
Cond 7100 e / 2(X)H
Cond 7100 e FF
Cond 7100 PA
Input for 2- or 4- electrode sensors
4-electrode 0.2 µS · c ... 1000 mS · c (c= cell constant)
2-electrode 0.2 µS · c ... 200 mS · c
(Display range limited to 3500 mS)
Resistivity
Salinity
USP
4)
Measurement error
1,2,3)
00.00 ... 99.99 µS/cm
000.0 ... 999.9 µS/cm
0000 ... 9999 µS/cm
0.000 ... 9.999 mS/cm
00.00 ... 99.99 mS/cm
000.0 ... 999.9 mS/cm
0.000 ... 9.999 S/m
00.00 ... 99.99 S/m
00.00 ... 99.99 MOhm cm
0.0 ... 45.0 ‰ (0 ... 35 °C)
00.00 ... 99.99 µS/cm
<1 % meas. val + 0.4 µS · c (c = cell constant)
-01- NaCl
-02- HCl
-03- NaOH
-04- H
2
SO
4
-05- HNO
3
0.00 … 9.99 by wt % (0 … 60 °C)
0.00 … 9.99 by wt % (– 20 … 50 °C)
0.00 … 9.99 by wt % (0 … 100 °C)
0.00 … 9.99 by wt % (–17 … 110 °C)
0.00 … 9.99 by wt % (–17 … 50 °C)
Input of cell constant with simultaneous display of conductivity and temperature
Input of conductivity of calibration solution with simultaneous display of cell constant and temperature
4)
Product calibration
4)
Temperature probe adjustment
00.0050 … 19.9999 cm
–1
Water monitoring in the pharmaceutical industry
Polarization detection and monitoring of the cable capacitance (can be disabled)
Provides information on the condition of the sensor, evaluation of Sensocheck
Direct display of measured values from sensor for validation, resistance / temperature
Pt 100 / Pt 1000 / NTC 30 kOhm / NTC 8.55 kOhm
2-wire connection, adjustable
Measurement ranges
Pt 100 / Pt 1000
NTC 30 kOhm
NTC 8.55 kOhm
– 20.0 ... + 200.0 °C (– 4 ... + 392 °F)
– 20.0 ... + 150.0 °C (– 4 ... + 302 °F)
– 10.0 ... + 130.0 °C (+ 14 ... + 266 °F)
Resolution 0.1 °C / 1 °F
Measurement error
1,2,3)
< 0.5 K (<1 K with Pt100; <1K with NTC >100 °C)
Specifications «Advanced Line» transmitters for conductivity measurement
Temperature (OFF)
no compensation
compensation
*)
(Lin)
Linear characteristic 00.00 ... 19.99 % / K – 20 … 130 °C reference temperaure 25 °C
(NLF)
Natural waters to EN 27888
(nACL)
Ultrapure water with NaCl traces
0 … 36 °C
0 ... 120 °C
(HCL)
Ultrapure water with HCl traces
(nH3)
Ultrapure water with NH3 traces
0 ... 120 °C
0 ... 120 °C
*) User-defined
1) According to IEC 746 part 1, at nominal operating conditions
2) ±1 count
3) Plus sensor error
4) Not valid for Cond 7100 PA
9
10
Specifications «Advanced Line» transmitters for conductivity measurement
Transmitters
Cond input
Working ranges
Effective ranges
*)
Concentration determination
Sensor standardization
Permissible cell factor
Permissible transfer ratio
Permissible zero point deviation
Sensocheck
Sensoface
Sensor monitor
2-wire HART
FOUNDATION™ Fieldbus
Profibus PA
-01NaCI
-02- HCI
-03- NaOH
-04- H
2
SO
4
-05- HNO
3
-06- H
2
SO
4
-07- HCI
-08- HNO
3
-09- H
2
SO
4
-10- NaOH
Cond Ind 7100 e / 2(X)H
Cond Ind 7100 e FF
5)
Cond Ind 7100 PA input for inductive sensors METTLER TOLEDO
Conductivity 0.000 ... 9.999 mS/cm
00.00 ... 99.99 mS/cm
000.0 ... 999.9 mS/cm
Concentration
Salinity
Measurement error
1, 2, 3)
0000 ... 1999 mS/cm
0.000 ... 9.999 S/m
00.00 ... 99.99 S/m
00.00 ... 9.99 %, 10.0 … 100.0 %
0.0 ... 45.0 ‰ (0 ... 35 °C)
<1 % meas. val + 0.02 mS/cm
0 – 26 by wt % (0 °C)
0 – 18 by wt % (–20 °C) ... 0 – 18 by wt % (50 °C)
0 – 13 by wt % (0 °C)
0 – 26 by wt % (–17 °C) ... 0 – 37 by wt % (110 °C)
0 – 30 by wt % (–20 °C) ... 0 – 30 by wt % (50 °C)
94 – 99 by wt % (–17 °C)
22 – 39 by wt % (–20 °C)
35 – 96 by wt % (–20 °C)
28 – 77 by wt % (–17 °C)
15 – 50 by wt % (0 °C)
... 0 – 28 by wt % (100 °C)
... 0 – 24 by wt % (100 °C)
... 89 – 99 by wt % (115 °C)
... 22 – 39 by wt % (50 °C)
... 35 – 96 by wt % (50 °C)
... 39 – 88 by wt % (115 °C)
... 35 – 50 by wt % (100 °C)
Input of cell factor with simultaneous display of conductivity value and temperature
Input of conductivity value with simultaneous display of cell factor and temperature
Product calibration
4)
Zero point calibration
Temperature probe adjustment
00.100 … 19.999
01.00 … 199.99
± 0.5 mS/cm
Monitoring of sender coil and leads for short circuiting, and of the receiver coil for circuits (can be disabled)
Indicates sensor status (zero point, Sensocheck)
Display of direct measurement values for validation purpose (resistance / temperature)
Specifications
Temperature input
*)
«Advanced Line» transmitters for conductivity measurement
Pt 100 / Pt 1000 / NTC 30 kOhm / NTC 100 kOhm
2-wire connection, adjustable
Measurement ranges
Pt 100 / Pt 1000
NTC 100 kOhm
NTC 30 kOhm
Resolution
Measurement error
1, 2, 3)
–20.0 ... + 200.0 °C (– 4 ... + 392 °F)
–20.0 ... + 130.0 °C (– 4 ... + 266 °F)
–20.0 ... + 150.0 °C (– 4 ... + 302 °F)
0.1 °C / 1 °F
< 0.5 K (<1 K with Pt100; <1K with NTC >100 °C)
Temperature compensation
(reference
*)
temperature 25 °C)
(OFF)
Without
(LIN)
linear, 0.00 … 19.99 % / K
(NLF)
Natural waters to EN 27888
*) User-defined
1) According to IEC 746 part 1, at nominal operating conditions
2) ±1 count
3) Plus sensor error
4) Not valid for Cond Ind 7100 PA
5) Available Q2 2005
11
12
Terminal assignment
Transmitter pH 2100 e / 2(X)H, pH 2100 PA and pH 2100 e FF
«Advanced Line» Transmitter
WARNING
- EXPLOSION HAZARD - NO NOT DISCONNECT EQUIPMENT UNLESS POWER
HAS BEEN SWITCHED OFF OR THE AREA IS KNOWN TO BE NON-HAZARDOUS.
WARNING
- BONDING BETWEEN CONDUIT IS NOT AUTOMATIC AND MUST BE
PROVIDED AS PART OF THE INSTALLATION. SEE INSTALLATION INSTRUCITONS.
AVERTISSEMENT
- RISQUE D’EXPLOSION. AVANT DE DECONNECTER L’EQIPEMENT,
COUPER LE COURANT OU S’ASSURER QUE L’EMPLACEMENT EST DESIGNE.
NON DANGEREUX.
67159
11 10 9 8
7
6 5 4
2 1
IEC 1158-2
PROFI-
BUS-PA input shield observe grounding conditions
0032
ZELM 99 ATEX xxxx
II 2 (1) G EEx ia IIC T4
T – 20 to + 55
°
C
Electrical Data see Type Examination Certificate
00000
Terminal assignment
Transmitter O
2
4100 e / 2(X)H, O
2
4100 PA and O
2
4100 e FF
«Advanced Line» Transmitter
13
14
Terminal assignment
Transmitter Cond 7100 e / 2(X)H, Cond 7100 PA and 7100 e FF
«Advanced Line» Transmitter
Terminal assignment
Transmitter Cond Ind 7100 e / 2(X)H, Cond Ind 7100 PA and Cond Ind 7100 e FF
«Advanced Line» Transmitter
15
16
General specifications «Advanced Line» Transmitter
Loop current
4)
HART communication
Supply voltage
Characteristic
Supply voltage
Overrange
*)
Output filter
*)
Meas. error
1)
Start/End of scale
*)
4 ... 20 mA (3.8 ... 20.5 mA), floating
Linear
12 ... 30 V I max
= 100 mA, P max
= 0.8 W
22 mA in the case of error messages
PT
1
-filter, filter time constant 0 ... 120 s
< 0.3 % of current value +0.05 mA as desired within measuring range
Digital communication via FSK modulation of the loop current, reading of device identification, measured values, status and messages reading and writing parameters, starting product calibration, signaling configuration amendment according to FDA 21 CFR Part 11.
Profibus communication
Protocol
Interface Profile
Supply voltage
Current consumption
Physical interface
Max. current in case of fault
Limit 1 and 2
Profibus PA via segment coupler / link to PLC
Profile for Analyzers Version 3.0 (PNO directive)
FISCO
≤
17.5 V
(trapezoidal or rectangular characteristic)
≤
24 V (linear characteristic) pH 2100 PA < 12.7 mA, O
2
4100 PA < 13.3 mA
Cond / Cond Ind 7100 PA < 16.0 mA according to EN 61158-2
(FDE) < 17.6 mA
Cyclical, discrete signal (DI) via Profibus, user-defined for the process variables.
FF communication
Physical interface
Address range
Mode of operation
To EN 61 158-2 (IEC 1158-2)
017 ... 246, Factory setting: 026
Bus-powered device with constant current
Supply voltage consumption
FISCO
≤
17.5 V (trapezoidal or rectangular characteristic)
≤
24 V (linear characteristic) pH 2100 e FF < 12.7 mA Current consumption
O
2
4100 e FF < 12.2 mA
Cond 7100 e FF < 16 mA
Cond Ind 7100 e FF < 16.1 mA
Max. current in case of fault (FDE)
Certified to ITK 4.6
Channel definition pH
O
2 pH 2100 e FF < 21.4 mA
O
2
4100 e FF < 17.6 mA
Cond 7100 e FF < 17.6 mA
Cond Ind 7100 e FF < 21.8 mA
1 resource block
1 transducer block
3 AI function blocks pH, ORP, temperature, R glass
, R ref
, asymmetry potential, slope
O
2 saturation, O
2 concentration, temperature, zero, slope, volume concentration in gas
Cond Conductivity, resistivity, concentration, salinity, temperature, cell constant
Cond Ind Conductivity, concentration, salinity, temperature, cell factor
General specifications «Advanced Line» Transmitter
Device description (DD)
FOUNDATION™ Fieldbus DD for AMS
Profibus PA DD for SIMATIC PDM
HART DD for AMS and SIMATIC PDM
Display
LC display
Main display
Secondary display
Sensoface
Mode indicators
7-segment with icons
Character height 17 mm, unit symbols 10 mm
Character height 10 mm, unit symbols 7 mm
3 status indicators
(friendly, neutral, sad smiley)
5 mode indicators “meas”, “cal”, “alarm”,
“digital communication”, “config”
18 further icons for configuration and messages
Red LED in case of alarm or HOLD, user-defined Alarm indication
5 keys [cal] [conf] [ ] [ ] [enter]
Keypad
Service functions
Device self-test
Display test
Last error
Sensor monitor
Automatic memory test (RAM, ROM, EEPROM)
Display of all segments
Display of last error occurred
Display of direct, uncorrected sensor
(resistance / temperature)
Data retention
Explosion protection
Parameters and calibration data >10 years (EEPROM)
EMC DIN EN 61326
Emitted interference
Immunity to interference
Class B; Class A: for > 60 V DC
Industrial sector
X100/2XH
X100 PA
X100 e FF
ATEX II 2 (1) G EEx ib (ia) IIC T6
FM IS, CIass 1, Div 1, Group A, B, C, D T4
NI, Class 1, Div 2, Group A, B, C, D T4
ATEX II 2 (I) G EEx ia IIC T4
ATEX II 2 (1) G EEx ia IIC T4
FM IS, Class 1, Div 1, Group A, B, C, D T4
NI, Class 1, Div 1, Group A, B, C, D T4
Nominal operation conditions
Enclosure
Ambient temperature
Transport / Storage temp.
–20 … + 55 °C
–20 … + 70 °C
Molded enclosure made of PBT (polybutylene terephthalat)
Color Bluish gray RAL 7031
Dimensions
Protection
Cable glands
Pipe mounting Ø 40 ... 60 mm, 30 ... 45 mm
Panel mounting cutout to DIN 43 700,
Sealed against panel
H 144 mm, W 144 mm, D 105 mm
IP 65 / NEMA 4X
3 breakthroughs for cable glands M20x1.5
2 breakthroughs for NPT
1
/
2
" or Rigid Metallic Conduit
*)
User-defined
1)
According to IEC 746 part 1, at nominal operating conditions
2)
± 1 count
3)
Plus sensor error
4)
Not available for Profibus ®
4)
Fieldbus versions
PA and FOUNDATION™
17
18
Ordering information
Description
2-wire instruments
Transmitter pH 2100 e / 2H
Transmitter pH 2100 e / 2XH
Transmitter pH 2100 e FF
Transmitter pH 2100 PA
Transmitter O
2
4100 e / 2H
Transmitter O
2
4100 e / 2XH
Transmitter O
2
4100 e FF
Transmitter O
2
4100 PA
Transmitter Cond 7100 e / 2H
Transmitter Cond 7100 e / 2XH
Transmitter Cond 7100 e FF
Transmitter Cond 7100 PA
Transmitter Cond Ind 7100 e / 2H
Transmitter Cond Ind 7100 e / 2XH
Transmitter Cond Ind 7100 e FF
Transmitter Cond I 7100 PA
Mounting accessories
Bracket kit
Panel-mount kit
Protective hood
1)
Available Q2 2005
Designation
pH 2100 e / 2H pH 2100 e / 2XH pH 2100 e FF pH 2100 PA
O
2
4100 e / 2H
O
2
4100 e / 2XH
O
2
4100 e FF
O
2
4100 PA
Cond 7100 e / 2H
Cond 7100 e / 2XH
Cond 7100 e FF
Cond 7100 PA
Cond Ind 7100 e / 2H
Cond Ind 7100 e / 2XH
Cond Ind 7100 e FF
Cond Ind 7100 PA
«Advanced Line» Transmitter
Order no.
52 120 724
52 120 758
52 121 245
52 121 042
52 121 215
52 121 168
52 121 246
52 121 091
52 120 903
52 120 905
52 121 247
52 121 047
52 120 908
52 120 910
52 121 248
1)
52 121 048
ZU 0274
ZU 0275
ZU 0276
52 120 741
52 120 740
52 120 739
Notes
19
METTLER TOLEDO Market Organizations
Sales and Service:
Australia
Mettler-Toledo Ltd.
220 Turner Street
Port Melbourne
AUS - 3207 Victoria
Phone +61 31300 659 761
Fax +61 3 9645 3935 e-mail [email protected]
Austria
Mettler-Toledo GmbH
Südrandstrasse 17
AT - 1230 Wien
Phone +43 1 604 19 80
Fax +43 1 604 28 80 e-mail [email protected]
Brazil
Mettler-Toledo Ind. e Com. Ltda.
Alameda Araguaia
451 - Alphaville
BR - 06455-000 Barueri / SP
Phone +55 11 4166 74 00
Fax +55 11 4166 74 01 e-mail [email protected]
China
Mettler-Toledo Instruments
(Shanghai) Co. Ltd.
589 Gui Ping Road
Cao He Jing
CN - 200233 Shanghai
Phone +86 21 64 85 04 35
Fax +86 21 64 85 33 51 e-mail [email protected]
Croatia
Mettler-Toledo d.o.o.
Mandlova 3
HR - 10000 Zagreb
Phone +385 1 292 06 33
Fax +385 1 295 81 40 e-mail [email protected]
Czech Republic
Mettler-Toledo spol s.r.o.
Trebohosticka 2283 / 2
CZ - 100 00 Praha 10
Phone +420 2 72 123 150
Fax +420 2 72 123 170 e-mail [email protected]
Denmark
Mettler-Toledo A/S
Naverland 8
DK - 2600 Glostrup
Phone +45 43 27 08 00
Fax +45 43 27 08 28 e-mail [email protected]
France
Mettler-Toledo
Analyse Industrielle Sàrl
30, Boulevard de Douaumont
BP 949
F - 75829 Paris Cedex 17
Phone +33 1 47 37 06 00
Fax +33 1 47 37 46 26 e-mail [email protected]
Germany
Mettler-Toledo GmbH
Prozeßanalytik
Ockerweg 3
D - 35396 Gießen
Phone +49 641 507 333
Fax +49 641 507 397 e-mail [email protected]
Great Britain
Mettler-Toledo LTD
64 Boston Road, Beaumont Leys
GB - LE4 1AW Leicester
Phone +44 116 235 7070
Fax +44 116 236 5500 e-mail [email protected]
Hungary
Mettler-Toledo Kereskedelmi KFT
Teve u. 41
HU - 1139 Budapest
Phone +36 1 288 40 40
Fax +36 1 288 40 50 e-mail [email protected]
India
Mettler-Toledo India Private Limited
Amar Hill, Saki Vihar Road
Powai
IN - 400 072 Mumbai
Phone +91 22 2857 0808
Fax +91 22 2857 5071 e-mail [email protected]
Italy
Mettler-Toledo S.p.A.
Via Vialba 42
I - 20026 Novate Milanese
Phone +39 02 333 321
Fax e-mail
+39 02 356 2973 [email protected]
Japan
Mettler-Toledo K.K.
Process Division
5F Tokyo Ryutsu Center, Annex B
6-1-1 Heiwajima, Ohta-ku
JP - 143-0006 Tokyo
Phone +81 3 5762 07 06
Fax +81 3 5762 09 71 e-mail [email protected]
Malaysia
Mettler-Toledo (M) Sdn Bhd
Bangunan Electroscon Holding
Lot 8 Jalan Astaka U8 / 84
Seksyen U8, Bukit Jelutong
MY - 40150 Shah Alam Selangor
Malaysia
Phone +60 3 78 45 57 73
Fax +60 3 78 45 87 73 e-mail [email protected]
Mexico
Mettler-Toledo S.A. de C.V.
Pino No. 350, Col. Sta.
MA. Insurgentes, Col Atlampa
MX - 06450 México D.F.
Phone +52 55 55 47 57 00
Fax +52 55 55 41 22 28 e-mail [email protected]
Poland
Mettler-Toledo (Poland) Sp.z.o.o.
ul. Poleczki 21
PL - 02-822 Warszawa
Phone +48 22 545 06 80
Fax +48 22 545 06 88 e-mail [email protected]
Russia
Mettler-Toledo Vostok ZAO
Sretenskij Bulvar 6/1
Office 6
RU - 101000 Moscow
Phone +7 095 921 92 11
Fax +7 095 921 63 53
+7 095 921 78 68 e-mail [email protected]
Singapore
Mettler-Toledo (S) Pte. Ltd.
Block 28
Ayer Rajah Crescent # 05-01
SG - 139959 Singapore
Phone +65 6890 00 11
Fax +65 6890 00 12
+65 6890 00 13 e-mail [email protected]
Slovakia
Mettler-Toledo s.r.o.
Bulharska 61
SK - 82104 Bratislava
Phone +421 243 42 74 96
Fax +421 243 33 71 90 e-mail [email protected]
Slovenia
Mettler-Toledo d.o.o.
Peske 12
SI - 1236 Trzin
Phone +386 1 530 80 50
Fax +386 1 562 17 89 e-mail [email protected]
South Korea
Mettler-Toledo (Korea) Ltd.
Yeil Building 1 & 2 F
124-5, YangJe-Dong
SeCho-Ku
KR - 137-130 Seoul
Phone +82 2 3498 3500
Fax +82 2 3498 3556 e-mail [email protected]
Spain
Mettler-Toledo S.A.E.
C/ Miguel Hernández, 69-71
ES - 08908 L’Hospitalet de Llobregat
(Barcelona)
Phone +34 93 223 76 00
Fax +34 93 223 76 01 e-mail [email protected]
Sweden
Mettler-Toledo AB
Virkesvägen 10
Box 92161
SE - 12008 Stockholm
Phone +46 8 702 50 00
Fax +46 8 642 45 62 e-mail [email protected]
Switzerland
Mettler-Toledo (Schweiz) AG
Im Langacher
Postfach
CH - 8606 Greifensee
Phone +41 44 944 45 45
Fax +41 44 944 45 10 e-mail [email protected]
Thailand
Mettler-Toledo (Thailand) Ltd.
272 Soi Soonvijai 4
Rama 9 Rd., Bangkapi
Huay Kwang
TH - 10320 Bangkok
Phone +66 2 723 03 00
Fax +66 2 719 64 79 e-mail [email protected]
USA / Canada
Mettler-Toledo Ingold, Inc.
36 Middlesex Turnpike
Bedford, MA 01730, USA
Phone +1 781 301 8800
Toll free +1 800 352 8763
Fax +1 781 271 0681 e-mail [email protected]
ISO
9001 certified
ISO
14001 certified
Management System certified according to
ISO 9001 / ISO 14001
Subject to technical changes.
© Mettler-Toledo GmbH, Process Analytics
01/05 Printed in Switzerland. 52 121 226
Mettler-Toledo GmbH, Process Analytics
Industrie Nord, CH - 8902 Urdorf
Phone + 41 44 736 22 11, Fax +41 44 736 26 36
www.mtpro.com
advertisement