Motic BA310 Pol Instruction manual


Add to my manuals
30 Pages

advertisement

Motic BA310 Pol Instruction manual | Manualzz
BA310 Pol
Polarizing Microscope
Instructions Manual
WWW.MOTIC.COM
MOTIC INCORPORATION LTD.
We are constantly endeavouring to improve our instruments and to adapt them to the requirements of
modern research techniques and testing methods. This involves modification to the mechanical
structure and optical design of our instruments.
Therefore, all descriptions and illustrations in this instruction manual, including all specifications are
subject to change without notice.
Although every effort has been made to ensure the accuracy of this instruction manual, if you note any
points that are unclear or inaccurate, please contact Motic agency or our Technical Service directly.
0
Introduction
Polarizing microscopes are used in the study of thin sections of minerals and rocks as well as other
anisotropic materials (fibers, plastics, etc.). The optical system is similar to that of modern compound
microscopes.
The most distinctive features of a polarizing microscope are the rotatable stage, polarizer and analyzer,
Bertrand lens system as well as mica plate, gypsum plate and quartz wedge. This special equipment
allows the evaluation of properties and characteristics of materials that cannot be measured by other
microscope methods.
1
Table of contents
1. Nomenclature
5
2. Setting Up The Instrument
9
3. Assembling the microscope
10
3.1
Verifying input voltage
10
3.2
Halogen lamp
11
3.3
Specimen clip
11
3.4
Attachable mechanical stage (optional)
11
3.5
Objectives
11
3.6
Condenser
11
3.7
Intermediate tube
12
3.8
Analyser slider
12
3.9
Compensators
12
3.10 Eyepiece tube
12
3.11 Eyepieces
12
3.12 Filters
13
3.13 Power cord
13
3.14 Epi-illuminator
4. Microscopy
14
Manipulation of each component
4.1
Coarse and fine focusing
14
A. Coarse focus torque adjustment
14
B. Coarse focus lock
14
4.2
Stage upper limit stop adjustment
15
4.3
Beam split slider
15
4.4
Interpupillary distance adjustment
15
4.5
Diopter adjustment
16
4.6
Centering the condenser
16
4.7
Use of aperture diaphragm
17
4.8
Use of field diaphragm
17
4.9
Brightness and contrast adjustment
18
4.10 Orientation of polarizer and analyzer
18
4.11 Focusing and centering the bertrand lens
19
2
4.12 Compensators
20
A. 1/4λ-plate
20
B. 1λ-plate
20
C. Quartz Wedge
20
4.13 Centering the Objectives
21
4.14
Epi-illuminator
5. Photomicrographic procedure
22
6. Terminology
23
7. Troubleshooting table
25
8. Care and maintenance
27
9. Warning labels
29
3
1. Nomenclature
BA310 EPI-Pol
4
BA310 EPI-Pol
5
6
2. Setting up the Instrument
Avoid placing the instrument in locations exposed to direct sunlight, dust, vibration, high temperature,
high humidity and where it is difficult to unplug the power supply cord.
Operating environment
 Indoor use
 Altitude: Max 2000 meters
 Ambient temperature: 15°C to 35°C
 Maximum relative humidity: 75% for temperature up to 31ºC decreasing linearly to 50% relative
humidity at 40ºC
 Supply voltage fluctuations: Not to exceed ±10% of the normal voltage
 Pollution degree: 2 (in according with IEC60664)
 Installation / Overvoltage category: 2 (in according with IEC60664)
 Air Pressure of 75kPa to 106kPa
 Avoid frost, dew, percolating water, and rain
7
3. Assembling the microscope
3.1 Verifying input voltage
 The automatic voltage selection works with a broad range of settings. However, always use a power
cord that is rated for the voltage used in your area and that has been approved to meet local safety
standards. Using the wrong power cord could cause fire or equipment damage.
 When using an extension cord, only use a power supply cord with a protective earth (PE) wire.
 In order to prevent electric shock, always turn the switch on the power supply off before connecting
the power cord.
 Electrical Specifications:
Halogen
Input: 90-240V~, 80VA, 50-60Hz
Lamp: 6V
Fuse: 250V
30W Halogen
T2.5A (If the original fuse is blown, please replace with specified fuse)
8
3.2 Halogen lamp
 Tungsten-halogen lamps operate at very high temperatures and may cause serious burn injuries if
handled while hot.
 When replacing these lamps, always allow them to cool down before removing them from the lamp
socket.
 Avoid handing the bulb directly because fingerprints will be burned into the glass, often initiating
premature lamp failure.
 The Halogen lamp bulb comes packed in protective plastic bags to avoid handling problems.
 Use a pair of scissors to cut the bag near the tungsten pins, insert the lamp into its holder while it still
remains in the bag.
 Remove the bag, when the lamp is properly positioned in the lamp socket.
3.3 Specimen clip
 Insert two specimen clips into the holes on the circular graduated stage surface.
3.4 Attachable mechanical stage (optional)
 Mount the attachable mechanical stage onto the rotating stage, while inserting the two positioning
pins of the mechanical stage into the holes of the stage surface.
 Tighten the clamp screws.
3.5 Objectives
 Lower the stage completely.
 Screw the objectives into the revolving nosepiece starting with the reference position (without
centering option) so that a clockwise rotation of the nosepiece brings the next higher magnification
objective into position.
Note: The BA310 Polarizing microscope allows you to center three objective positions in relation to the
reference objective position.
3.6 Condenser
 Raise the stage by turning the coarse focus knob clockwise.
 Lower the condenser carrier by turning the condenser focus knob anti-clockwise.
 Insert the condenser in the dovetail mount with aperture scale facing the front.
 Secure the condenser with the clamp screw.
 Turn the condenser focus knob to raise the condenser as far as it will go.
9
3.7 Intermediate tube
 To mount the intermediate tube, start loosening the intermediate tube clamp screw on the
microscope arm.
 Insert the intermediate tube into the round dovetail mount on the microscope arm.
 The orientation pin on the lower side of the intermediate tube has to fit to the receiving groove on
the microscope arm.
 Secure in position with the clamp screw.
A built-in depolarizer unit inside the intermediate tube provides equalized colours in the eyepiece tube
and the photo port. (See Microscope Terminology)
3.8 Analyser slider
•
Unscrew the analyser slider knob located on the side of the slider.
 Insert the slider into the slot of the intermediate tube with the analyser scale
positioned on the right hand side.
 Secure the slider by relocating the slider knob in the related thread hole.
3.9 Compensators
•
Compensators will be inserted into the compensator slot of the intermediate tube.
•
The compensator slot is oriented at 45º to the cross-lines of the eyepiece.
3.10 Eyepiece tube
•
Loosen the eyepiece tube clamp screw. Insert the eyepiece tube on the dovetail mount on top of
the intermediate tube.
 Make sure the location pin on the lower side of the eyepiece tube matches the pin groove of the
intermediate tube.
•
Secure the eyepiece tube in position with the clamp screw.
3.11 Eyepieces
•
Place the eyepieces into the sleeves of the binocular eyepiece tube with a slightly turning move.
•
The sleeve of the right eyepiece tube has positioning slots for either 90° and 45° orientation
•
Install the correct eyepiece by aligning the positioning pin of the eyepiece with positioning slot of the
eyepiece sleeve.
•
Slide the rubber eyecups onto each eyepiece.
10
3.12 Filters
•
Place the chosen filter in the filter holder located above the field lens by removing the filter holder
ring. Avoid polluting filter and lens surfaces with dust, dirt or finger prints.
Filter options:
Filter
Function
ND2 (Transmission=50%)
ND4 (T=25%)
For brightness adjustment in photomicrography
ND16 (T=6.25%)
Blue Filter (Colour Balancing Filter)
Green Interference (546nm)
For routine microscopy and photomicrography
For retardation measurement and contrast
adjustment
3.13 Power cord
•
Connect the socket of the of the power cord to the AC inlet on the rear of the microscope. Plug in
the other end of the cord to an AC outlet with ground conductor.
3.14 Epi-illuminator
•
Take Epi-illuminator instruction manual for reference.
11
4. Microscopy
Manipulation of each component
4.1 Coarse and fine focusing
•
Focusing is carried out with the coarse and fine focus knobs at the left and right side of the
microscope stand.
•
The direction of vertical movement of the stage corresponds to the turning direction of the focus
knobs.
•
One complete rotation of the fine focus knob moves the stage 0.2mm. The graduation on the fine
focus knob is 2 microns.
Never attempt either of the following actions, since doing so will damage the focusing mechanism:
Rotate the left or right knob while holding the other.
Turning the coarse and fine focus knobs further than their limit.
A. Coarse focus torque adjustment
•
To increase the torque, turn the torque adjustment ring located behind the left-hand coarse focus
knob in the direction indicated by the arrow. To reduce the torque, turn the ring in the opposite
direction.
B. Coarse focus knob lock  Coarse focus Quick Stop
•
The coarse focus knob lock  Coarse focus quick stop (on right side of microscope) can fix the upper
stage position at any height.
•
With the specimen in focus, turn the handle to fix the knob.
•
When the coarse focus knob lock  Coarse focus quick stop is in position, the stage cannot be
raised from that position.
12
4.2 Stage upper limit stop adjustment:
(The upper Stage Limit is preset at the factory; please only adjust if necessary)
•
The Stage Upper Limit stop marks the uppermost stage position. This is to prevent damage of the
objectives while focussing. If the correct preset has been changed accidently during transport, please
perform the following procedure:
- Loosen the Stage upper Limit Stop.
- Focus the specimen carefully.
- Fix the Stage upper Limit Stop.
 When the stage upper limit stop is in position, the Stage cannot be raised above that position.
However, the fine focus knob can be moved without raising the stage.
4.3 Beam split slider
•
The beam split slider of the trinocular eyepiece tube can be used to select the light distribution
between the eyepiece tube and the vertical phototube.
•
When the slider is pushed in, 100% of the light enters the eyepieces. When the prism slider is pulled
out, 100% of light will enter the phototube.
4.4 Interpupillary distance adjustment
•
Before adjusting the interpupillary distance, bring a specimen into focus using the 10x objective.
•
Adjust the interpupillary distance so that both the right and left field of view become one.
•
This adjustment will enable the user to observe the specimen with both eyes.
13
4.5 Diopter adjustment
•
Every human eye is different, so only the adjustment of the eyepieces will bring the microscope to its
best performance.
•
Set the diopter on both eyepieces to the “0” position.
•
Change to 10x magnification and focus the image of the specimen with one eye only.
•
Use the eye which is most convenient for first focussing.
•
When the best focus position is reached, close this eye and use the other eye for the following steps.
•
Correct the focus for the second eye by using only the Diopter Ring, do not use the coarse / fine
focusing knob!
•
Change to a higher magnification to verify the result and if necessary repeat the procedure.
Higher magnification objectives do have a smaller depth of focus, so the adjustment will be
more precise.
•
Keep this final diopter position for all magnification / lenses. The diopter position for each user can
be read from the scale, so it can easily be reset.
4.6 Centering the condenser
•
Fully open the field of view diaphragm and condenser aperture diaphragm.
•
Set the specimen on the stage with the cover glass facing up.
•
Bring the specimen image into focus, using the 10X objective.
•
Switch in the condenser head.
•
Close the field diaphragm by 50%.
•
Turn the condenser focus knob to bring the field diaphragm image into focus of the specimen
plane. The border of the light spot should be as sharp as possible.
•
By using the condenser centering screws, bring the light spot into the middle of the field.
•
For each objective this procedure should be repeated. Finally open the field diaphragm so that it is
just outside the field of view.
14
4.7 Use of aperture diaphragm
•
The condenser aperture diaphragm is provided for adjusting the numerical aperture (N.A.) of the
illumination system of the microscope. It determines the resolution of the image, contrast, depth of
focus and brightness.
•
Stopping down will lower the resolution and brightness but increase the contrast and depth of focus.
•
An image with appropriate contrast in most cases can be obtained with an aperture diaphragm
closed down to 2/3 of the maximum value.
•
To get this aperture diaphragm position:
- adjust the condenser aperture diaphragm ring referring to the condenser aperture scale, or
- by observing the diaphragm image visible on the exit pupil inside the eyepiece tube (the eyepiece
has to be removed for that), or
- by using a centering telescope after removing one of the eyepieces and focusing on the aperture
diaphragm.
•
When swinging-out the top lens of the condenser for low magnification objectives, open the
condenser aperture diaphragm completely.
4.8 Use of field diaphragm
•
The field diaphragm determines the illuminated area on the specimen. Turning the field diaphragm
ring changes the size of the field diaphragm. For normal observation, the diaphragm is set slightly
larger than the field of view. If a larger than required area is illuminated, stray light will enter the field
of view. This will create a flare in the image and lower the contrast.
•
The thickness of the glass slide must be 1mm, otherwise the field diaphragm may not be focused on
the specimen plane.
•
When the condenser top lens is swung out of the optical path, fully open the field diaphragm, as
the N.A. of the illuminating system will be reduced if the diaphragm is excessively stopped down.
15
4.9 Brightness and contrast adjustment
•
Neutral density filters are used for brightness adjustment in routine microscopy and
photomicrography without affecting the colour temperature.
•
The Green interference filter (546nm) is used in combination with the quartz wedge.
4.10 Orientation of polarizer and analyzer
•
Push in the brightfield position of the analyzer slider
•
Focus on the specimen.
•
Take the specimen out of the optical path.
•
Bring the analyzer into the optical path.
•
Set the analyser to the “0” position by loosening the analyzer clamp screw and rotating the analyser
dial (the rotation angle of the analyser can be read in the range between 0° to 360° in increments of
0.1° by the Vernier scale).
Dark cross image is formed on
the exit pupil of the objective
•
Rotate the Bertrand lens turret to “B” position and bring the Bertrand lens in the optical path to enable
the exit pupil of the objective to be seen through the eyepiece. Rotate the polarizer so that a dark
cross image is formed on the exit pupil as shown in the figure above.
16
4.11 Focusing and centering the Bertrand lens
•
Rotate the Bertrand lens turret to “B” position and bring the Bertrand lens in the optical path.
•
Bring 40x Objective into optical path.
•
Use the Bertrand lens focus ring under the Bertrand lens turret to focus on the image of the
condenser aperture diaphragm that is stopped down to 70 – 80% of the numerical aperture of the
objective.
•
Use the Bertrand lens centering screws to bring the image of the condenser aperture to the center
of the field of view.
•
The centering procedure is the same as that for the condenser except that the condenser aperture
diaphragm image is used instead of the field diaphragm image.
4.12 Compensators
•
All polarizing microscopes are equipped with a compensator in the optical beam path between
polarizer and analyzer.
•
Compensators are sections of optically anisotropic material which, when inserted in the microscope
between crossed polarizers, produce a defined optical path difference of mutually perpendicular
plane-polarized light waves.
A. 1/4λ-plate - The 1/4λ-plate is also called mica plate and has an optical path difference around 140nm
(yellow light has λ = 580nm, therefore 1/4λ = 580/4 = 145nm).
17
B. 1λ-plate (sensitive tint or first-order red) - The 1λ-plate is a gypsum plate of a thickness to give1λ
optical path difference for green light of 550 nm. This wavelength is therefore extinguished with the
resulting interference colour having the typical tint of the first-order red/violet. This magenta colour is
sometimes named ‘sensitive tint’.
C. Quartz Wedge – This device has a range of 4 orders and is commonly employed for qualitative
retardation measurements of petrographic specimens or other birefringent materials whose
retardation value falls within the wedge limit.
Measuring Retardation from 1λ to 4 λ
•
Observing extinction position
■ Observe the position where the part of the specimen to be measured becomes darkest by rotating
the stage under crossed polars.
•
Observing subtraction position
■ Rotate the stage 45° to the diagonal position; here the specimen appears to be at it’s brightest.
Insert the appropriate compensator into the slot; verify the interference colour of the part to be
measured. Rotate the stage another 90° to get the associated colour. With those 2 observed
colours, the +/- position of the sample can be determined.
•
Measuring retardation
■ Slide the quartz wedge along the length of the slot, and the interference colour will change.
■ Stop sliding the quartz wedge where the dark band covers the part of the specimen to be
measured. Remove the specimen and compare the interference colour (at the same position
without the specimen) with the Interference Colour Chart to ascertain the amount of retardation.
■ Measuring the retardation or ascertaining by interference colour should be done with field of view
diaphragm stopped down to the part to be measured.
Note: With monochromatic light the coloured fringes are seen as alternating dark and bright bands.
18
4.13 Centering the objective
•
The BA310 Polarizing microscope allows for centering of three objectives in relation to a reference
objective.
•
Before centering the objectives, look through the binocular tube at the field of view, pick out an easily
recognizable target and move the target in the middle to the cross line by moving the sample.
•
Rotate the stage
•
The target now rotates in a circle while touching at one point the intersection of the cross lines.
Rotate the stage until the target is farest away from the intersection. By means of the centering
screws bring it half way to the intersection point.
Move the target again to the intersection point by moving the slide. Repeat the procedure until the
target rotates on the intersection point.
Rotating stage
4.14 EPI Illuminator
• Take Epi-illuminator instruction manual for reference.
19
5. Photomicrographic Procedure
•
To ensure vibration free operation, set the microscope on a sturdy vibration free table or a bench with
a vibration proof device.
•
Pull the beam split lever of the trinocular eyepiece tube all the way out to the limit, the ratio of light
entering the phototube will be 100%
•
To ensure optimal illumination, check the position and centering of the lamp and position of the
condenser.
•
Select a blue filter for routine application.
•
Adjustment of the field diaphragm is important for the purpose of limiting extraneous light that may
cause flare and lower the contrast. Stop down the diaphragm to achieve an illuminated area slightly
larger than that of the field of view.
•
A change to depth of focus, contrast and resolution of the image is attainable with an aperture setting
that is 2/3 of the objective N.A.
•
For photomicrographic procedures, refer to the manual of the specific camera being used.
Note: The correlation between the position of the polarizing plate and photomicrographic devices should
not be of any concern to the user, given that the intermediate tube contains a built-in depolarizer.
20
6. Terminology
Aperture, Numerical (N.A.)
Diaphragm, Condenser
The numerical aperture is an important factor
A diaphragm, which controls the effective size of
determining the resolution power of the
the condenser opening angle and thus the
condenser and objective. It is calculated by the
illumination aperture.
formula:
(N.A. = ηsin α), where η is the refractive index of
Diopter Adjustment
a medium (air, water, immersion oil etc.) between
The adjustment of the eyepiece of an instrument
the objective and the specimen, and α is half of
to provide accommodation for the eyesight
the maximum angle at which light enters the lens
differences of individual observers.
from a focused object point on the optical axis.
Epi-Illumination
Anisotropic
Illumination is accomplished by means of a
Quality of a transparent material having different
vertical illuminator placed above the objective.
refractive indices depending on the vibration
direction of the transmitted light.
Focusing Eyepiece
An eyepiece with a device for focusing a graticule
Axis - Optical
mounted within it and coinciding with the primary
A straight line joining the centers of curvature of
image plane.
lens surfaces.
Michel Levy Scale
Bertrand Lens
Colour chart-plotting thickness of the anisotropic
An intermediate lens, which transfers an image of
specimen, its birefringence (n1-n2) and its
the back focal plane of the objective into the
retardation in nanometers. Any one of the three
primary image plane. Mostly used for conscopic
variables can be determined if the other two are
observation in polarized light microscopy.
known.
Cover Glass Thickness
Micrometer: um
Transmitted light objectives are designed to
A metric unit of length measurement=
image specimens that are covered by a thin cover
1x10-6 meters or 0.000001 meters
glass (cover slip). The thickness of these small
glass pieces is now standardized at 0.17 mm for
most applications.
21
Conoscopic Observation
Orthoscopic Observation
The study of the back focal plane of the objective
This is the normal way of viewing an object
is called conoscopic as the observations are
through a microscope. With Koehler illumination,
associated with the cone of light furnished by the
the image of the field diaphragm is projected by
condenser and viewed by the objective. This
the condenser into the specimen plane.
study is accomplished by inserting a Bertrand
lens or by using a phase telescope.
Stage Micrometer
The graduated scale used as a standard on the
Crossed Polars
stage of a light microscope for calibrating an
The condition in which the vibration directions of
eyepiece micrometer.
polarizers and analysers are mutually
perpendicular.
x-axis
In a plane Cartesian coordinate system, the
Depolarizer
horizontal axis, or axis in the left to right direction.
Depolarizers change plane polarization into a mix
of polarization states, which
y-axis
resemble un-polarized light.
In a plane Cartesian coordinate system, the
vertical axis orthogonal to the x-axis.
Nanometer: nm
A metric unit of length measurement=
1x10-9 meters or 0.000000001 meter.
22
7. Troubleshooting Table
As you use your microscope, you may occasionally experience a problem.
The troubleshooting table below contains the majority of frequently encountered problems and the
possible causes.
Problem
Possible Cause
Analyser slider in intermediate position
Compensator slider in midway position
Bertrand lens in optical path
Condenser not mounted correctly
Condenser is not centered
Vignetting or uneven brightness in the field of
view or field of view only partially visible
Condenser is set too low
Condenser top lens not fully swung in or out
Field diaphragm closed too far
Aperture diaphragm closed too far
Revolving nosepiece not clicked into position
Trinocular eyepiece tube beam splitter lever in
intermediate position
Aperture diaphragm closed too far
Dust or dirt in the field of view
Condenser is set too low
Dust or dirt on specimen surface
Dust or dirt on field lens, filter
Condenser is set too low
Aperture diaphragm closed too far
No cover glass
Too thick or thin cover glass
Immersion oil not used on immersion procedure
Poor image (low contrast or resolution)
Uneven focus
Air bubbles in immersion oil
Specified immersion oil used not used
Immersion oil on dry objective
Greasy residue on eye lens
Incorrect illumination
Specimen holder not fixed securely on stage
Specimen not secured in position
23
Image tinged yellow
Focusing is not possible with high
magnification objectives
High magnification objectives strike the
specimen when changing over from low to
high magnification
No cohesion of binocular image
Lamp voltage is set too low
Blue filter is not in
Slide is upside down
Cover glass is too thick
Slide is upside down
Cover glass is too thick
Eyepiece diopter not adjusted
Magnification or field of view of left and right
eyepieces differ
Interpupillary distance not adjusted
Eyepiece diopter not adjusted
Interpupillary distance not adjusted
Eye strain or fatigue
Diopter adjustment not made
Field of view of left and right eyepiece differ
Inadequate illumination
Electrical
Problem
Possible Cause
Power supply not plugged in
Lamp does not light
Lamp not installed
Lamp burnt out
Inadequate brightness
Specified lamp not being used
Lamp blows out immediately
Specified lamp not being used
Connectors are not securely connected
Lamp flickers
Lamp near end of service life
Lamp not securely plugged into socket
24
8. Care and Maintenance
A. Do not disassemble
1. Disassembly may significantly affect the performance of the instrument, and may result in electric
shock or injury and will void the terms of the warranty.
2. Never attempt to dismantle any parts other than described in this manual. If you notice any
malfunction, contact your nearest Motic representative.
B. Cleaning the microscope
•
Do not use organic solvents such as ether; alcohol or paint thinner on painted surfaces or plastic
components. Doing so could result in discolouration of painted or plastic surfaces.
•
When cleaning lenses do not use any solvents other than absolute alcohol as, they may damage lens
bonding cement.
•
Do not use petroleum benzene when cleaning components such as filters or lenses.
•
Absolute alcohol and petroleum benzene are highly flammable. Keep away from open flames and
when turning the power switch on or off.
•
For stubborn dirt, dampen a piece of gauze with diluted neutral detergent and wipe lightly.
C. Disinfecting the microscope
•
Follow the standard procedures for your laboratory.
D. When not in use
•
When not in use, cover the instrument with vinyl dust cover and store in a place low in humidity where
mould is not likely to form.
•
Store the objectives, eyepieces and filters in a container or desiccator with drying agent.
•
Proper handling of the microscope will ensure years of trouble free service.
•
If repair become necessary, please contact your Motic agency or our Technical Service direct.
Note:
•
If equipment is used in a manner not specified by the manufacturer, the protection provided by the
equipment may be impaired.
•
To avoid getting wet, do not use the microscope near water.
25
E. Bulb replacement
The bulb and the lamphouse become very hot during and after a period of operation.
Risk of burn – Do not touch the bulb during or immediately after period of operation.
Make sure the bulb has cooled sufficiently before attempting to replace the bulb.
•
The applicable halogen bulbs are the 6V 30HAL hign-intensity bulb.
•
In order to prevent electric shock always turn the power switch off and unplug the power cord before
installing or replacing the bulb.
•
Place the microscope on its back and pull back the lamp cover plate.
•
Firmly insert the bulb into the socket pinholes until it reaches the limit. Be careful not to tilt the bulb
when mounting.
•
When installing the bulb, do not touch the glass surface of the bulb with bare fingers. Doing so will
cause fingerprints, grease, etc., to burn onto the bulb surface, reducing the illumination provided by
the bulb. If the surface is contaminated, wipe it clean using lens tissue.
•
Close lamp cover plate and secure until it snaps into position.
26
9. Warning labels
The following warning labels (or symbols) are found on the microscope, study the meaning of the
warning labels (or symbols) and always use the equipment in the safest possible manner.
Warning Label / Symbol
Explanation
Indicates that the surface becomes hot, and should
not be touched with bare hands.
Indicates that the main switch is ON.
Indicates that the main switch is OFF.
~
Indicates alternating current.
CAUTION! Risk of danger.
Please consult documentation in all cases where
this symbol is used.
The lamp and the lamphouse become very hot during and after a period of operation.
Risk of burn – Do not touch the lamp during or immediately after period of operation.
Make sure the lamp has cooled sufficiently before attempting to replace the lamp.
Don’t pick up from the bottom during equipment operation.
Proper handling of the microscope will ensure years of trouble free service.
If repair become necessary, please contact your Motic agency or our Technical Service directly.
27
NO.: 1300901108661
Motic Incorporation Ltd. (Hong Kong)
Rm 2907-8, Windsor House, 311 Gloucester Road, Causeway Bay, Hong Kong
Motic Instruments Inc. (Canada)
130-4611 Viking Way, Richmond, B.C., V6V 2K9, Canada
Motic Deutschland GmbH (Germany)
Christian-Kremp-Strasse 11, D-35578 Wetzlar, Germany
Tel: 852-2837 0888
Tel: 1-877-977 4717
Tel: 49-6441-210 010
Motic Spain, S.L. (Spain)
Fax: 852-2882 2792
Fax: 1-604-303 9043
Fax: 49-6441-210 0122
Polígon Industrial Les Corts, Camí del Mig, 112 08349 Cabrera de Mar, Barcelona Spain
Tel: 34-93-756 6286
Fax: 34-93-756 6287
Website: http://www.motic.com
E-mail: [email protected]
Motic China Group., Ltd. (China)
Motic Building, Torch Hi-Tech Industrial, Development Zone, Xiamen P.R.C.
Tel: 86-0592-562 7866
Fax: 86-0592-562 7855
© 2007-2008 Motic China Group Co., Ltd. All rights reserved. Motic is a registered trademark and service mark of Motic China Group Co., Ltd.
Microsoft Windows logo is a registered trademark of Microsoft Corporation. All other trademarks are the property of their respective owners.
Design Change: The manufacturer reserves the right to make changes in instrument design in accordance with scientific and mechanical progress, without
notice and without obligation.
Updated 2012.03.21