Watts CLX Online Chlorine Owner's Manual

Add to My manuals
37 Pages

advertisement

Watts CLX Online Chlorine Owner's Manual | Manualzz
OWNER’S MANUAL
CLX OnLine Residual
Chlorine Monitor
HF scientific, inc.
3170 Metro Parkway
Ft. Myers, FL 33916
Phone: 239-337-2116
Fax: 239-332-7643
EMail:[email protected]
Website: www.hfscientific.com
Catalog #24420 (9/07)
REV. 3.1
DECLARATION OF CONFORMITY
Application of Council Directive
Standards to Which Conformity is Declared:
Product Safety - Tested and passed CE EN61010-1: 1990 + A1: 1992 (73/32 EEC)
st
- Tested and passed ETL (tested to UL3111-1), 1 Edition, 1994,
w/Bulletin June 5, 1996
- Tested and passed ETLc (tested to CSA C22.2#1010.1-92)
Immunity – Tested and passed EN61326: 1997+A1: 1998
Including:
IEC61000-4-2
IEC 61000-4-3
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-8
IEC 61000-4-11
Emissions - Tested and passed EN61326: 1997+A1: 1998
Including:
CISPR 11 Radiated Emissions Class B
CISPR 11 Conducted Emissions Class B
EN 61000-3-2
EN 61000-3-3
Manufacturer’s Name:
HF scientific, inc.
Manufacturer’s Address:
3170 Metro Parkway, Fort Myers, Florida 33916-7597
Importer’s Name:
Importer’s Address:
Type of Equipment:
Chlorine Process Analyzer
Model No:
CLX
I, the undersigned, hereby declare that the equipment specified above conforms to the
above Directive and Standard
Place:
Fort Myers, Florida USA
(Signature)
Rowan T. Connelly, General Manager
Catalog #24420 (9/07)
REV. 3.1
Table of Contents
Section
Page
Specifications ....................................................................................................1
1.0
Overview ...........................................................................................................2
1.1
Unpacking and Inspection of the Instrument and Accessories .............2
1.2
The Display ...........................................................................................2
1.3
The Touch Pad ......................................................................................3
2.0
Safety.................................................................................................................4
2.1
Symbols Used In CLX ...........................................................................4
3.0
Theory of Operation ........................................................................................5
4.0
Installation and Commissioning ....................................................................6
4.1
Mounting and Site Selection..................................................................6
4.2
Plumbing ..............................................................................................7
4.3
Setting Flow Rate...................................................................................8
4.4
Electrical Connections ..........................................................................8
4.4.1 Power ........................................................................................9
4.4.2 RS-485 ...................................................................................10
4.4.3 Relays.......................................................................................10
4.4.4 4-20 mA ..................................................................................10
4.4.5 RS-485/4-20 mA cable Ferrite.................................................10
5.0
Operation .......................................................................................................11
5.1
Routine Measurement .........................................................................11
5.2
Security Access Feature.......................................................................12
6.0
Instrument Calibration ................................................................................13
6.1
Slope(Gain) Calibration Procedure .....................................................13
6.2
Zero(Offset) Calibration Procedure ....................................................14
6.3
Restore Factory Settings ......................................................................14
7.0
Instrument Configuration (CONFIG mode) ...............................................15
7.1
Setting the 4-20 mA .............................................................................15
7.2
Configuring the Error Level.................................................................16
7.3
Configuring the RS-485 Port ..............................................................17
7.4
Configuring the Alarms .......................................................................17
7.4.1 Alarm 1 ....................................................................................18
7.4.2 Alarm 2 ...................................................................................18
7.5
Enabling the Security Access...............................................................18
7.6
Extended Settings.................................................................................19
7.7
Speed of Response ...............................................................................19
Table of Contents (continued)
Section
Page
7.8
7.9
7.10
7.11
LCD Backlight Brightness...................................................................19
RS- 485 Parameters..............................................................................20
Cycle Time ..........................................................................................20
Saving Configuration Settings .............................................................20
8.0
Additional Features and Options ................................................................21
8.1
Backlit LCD .........................................................................................21
8.2
RS-485 Output .....................................................................................21
8.2.1 HF Online Communication......................................................21
8.2.2 Simple Communication ...........................................................21
8.2.3 Modbus Communication..........................................................22
8.3
Remote Panel Meter.............................................................................22
8.4
Desiccant Cartridge..............................................................................22
9.0
Troubleshooting .............................................................................................23
9.1
CLX Fault Detection............................................................................23
9.2
Clearing Faults .....................................................................................24
9.3
Diagnostic Chart .................................................................................24
9.4
Technical and Customer Assistance ...................................................24
10.0
Routine Maintenance.....................................................................................25
10.1 Maintenance Schedule ........................................................................25
10.2 Replacing or Installing the Reagents ...................................................28
10.3 Cuvette Change ...................................................................................28
10.4 Instrument Storage ..............................................................................29
10.5 Cleaning the CLX ...............................................................................29
11.0
Accessories and Replacement Parts List .....................................................30
12.0
Warranty .......................................................................................................31
Specifications
Measurement Range
0 – 10.00 mg/L (PPM)
Accuracy
±5% or ±0.03 mg/L (PPM) for range of 0-6 mg/L(PPM)
Resolution
0.01 mg/L (PPM)
Cycle Time
Adjustable: 90 seconds to 10 minutes (defaults to 2.5 minutes)
Display
Multi-Line Liquid Crystal Backlit Display
Alarms
Two Programmable, 120-240VAC 2A Form C Relay
Analog Output
Powered 4-20 mA, 600 Ω drive, isolated
Communications Port
Bi-directional RS-485 with Modbus
Maximum Water Pressure
Integral pressure regulator rated 1380kPa (200 PSI.)
Flow Rate
200 ml/min. minimum flow
Operating Temperature
5°C – 40°C (41°F – 104°F)
Wetted Materials
Nylon, Borosilicate Glass, Silicon, Polypropylene, Stainless Steel, Brass
Sample Temperature Range
5°C – 40°C (41°F – 104°F)
Power Supply
100 – 240 VAC, 47 – 63 Hz, 150VA
Insulation Rating
Double Insulated, Pollution Degree 2, Overvoltage Category II
Environmental Conditions
Not recommended for outdoor use.
Altitude up to 2000 meters
Up to 95 % RH (non-condensing)
Enclosure Rating
Designed to meet IP 66 /NEMA 4X
Regulatory Compliance
CE Approved, ETL listed to UL 3111-1 &
And Certifications
ETL Certified to CSA 22.2 No. 1010-1-92
Shipping Weight
2.5 kg (5.5 lbs.)
Shipping Dimensions
406 mm X 406 mm X 229 mm (16”X 16” X 9”)
Warranty
1 Year from date of shipment
CLX (9/07)
REV 3.1
Page 1
1.0
Overview
The CLX Online Chlorine Analyzer allows for the reading of chlorine levels of process
water on-line. The CLX has been designed to meet the design criteria specified by
Standard Methods for the Examination of Water and Wastewater (20th Edition) Method
4500-Cl G. DPD Colorimetric Method. The CLX uses a 515nm LED as the measurement
light source.
1.1 Unpacking and Inspection of the Instrument and Accessories
The table below indicates the items in the shipment.
Item
Quantity
CLX Analyzer
1
Instruction Manual
1
Mounting Kit
1
Tubing/Cuvette Kit
1
In-Line Strainer
1
Remove the instrument from the packing carton. Carefully inspect all items to ensure that
no visible damage has occurred during shipment. If the items received do not match the
order, please immediately contact the local distributor or the HF scientific, inc. Customer
Service Department.
1.2 The Display
Figure 1 illustrates all the items that can appear on the display. The upper row of the
display (1) is used for reporting the chlorine levels and to provide user guidance in the
customer setting routine. The lower row of the display (2) is used to communicate error
messages (message queue) and provide user guidance. The display has two icons (3) that
are used to indicate the use of access code and offset mode. In addition, mode arrows (4)
are used to indicate the current instrument operating mode; AUTO (normal operation),
CAL (calibration) and CONFIG (configuration).
Figure 1: Display used in the instrument.
All items used on the display are shown in this figure.
CLX (9/07)
REV 3.1
Page 2
1.3 The Touch Pad
Figure 2 illustrates the touch pad. The touch pad has six buttons: PRIME, SERVICE,
MODE/EXIT, , S and T
The MODE/EXIT button is used to cycle between the three operational modes of the
instrument: CAL, CONFIG, and AUTO (Measurement) mode. The button enters the
option or mode that is highlighted or chosen. The S and T buttons are used to change
settings.
The PRIME and SERVICE buttons are dedicated controls. The PRIME will start 75
reagent pump pulses to prime the tubing after a change or addition of reagent bottles. The
SERVICE button will drain the instrument and hold all operations until either the
SEVICE button is pushed again or the power is reset. This button should be used while
changing the tubing, the measurement cuvette or reagent bottles.
Figure 2: The CLX touch pad.
CLX (9/07)
REV 3.1
Page 3
2.0
Safety
This manual contains basic instructions that must be followed during the commissioning,
operation, care and maintenance of the instrument. The safety protection provided by this
equipment may be impaired if it is commissioned and/or used in a manner not described in
this manual. Consequently, all responsible personnel must read this manual prior to
working with this instrument.
In certain instances “Notes”, or helpful hints, have been highlighted to give further
clarification to the instructions. Refer to the Table of Contents to easily find specific
topics and to learn about unfamiliar terms.
2.1 Symbols Used In CLX
Standard IEC symbols are used on the high voltage cover.
ISO 3864, No. B.3.6 Caution, risk of electric shock.
This symbol indicates that hazardous voltages may be present under this
cover
ISO 3864, No.B3.1 Caution refer to accompanying documents.
This symbol is reminding you to read the sections in the manual referring
to the electrical connections, and potential hazards.
CLX (9/07)
REV 3.1
Page 4
3.0
Theory of Operation
The CLX has two solenoid valves, one for sample water (FLOW) and one for draining of
the cuvette (PURGE). A third solenoid, along with four check valves forms a reagent
pump. Sample water flow is controlled by the FLOW solenoid valve. The PURGE
solenoid valve is used to empty the measurement cuvette.
The measurement chamber consists of a sample inlet, a purge drain, and an overflow. The
reagent is added from the top. A green LED provides the 515 nm source lamp, a red LED
is used for sample level and flow measurement. A single detector is located 180 º from the
green LED. A replaceable glass cuvette separates the LEDs from the detector and
maintains the measurement path length. Sample water flows in the inlet at the bottom,
through the measurement cuvette and out through overflow drain. This flow is used to
both fill the cuvette and flush the system.
The reagents consist of two replaceable bottles. One bottle has a buffer to control the pH;
the second has an indicator that contains the DPD, which produces color when chlorine is
present in the sample. The degree of color is dependant on the amount of chlorine in the
sample water.
The measurement chamber is open to view operations. A white LED backlights the
chamber for a clearer view. The white LED will flash to attract attention in the case of a
warning or failure. Most warnings and failures are displayed on the screen.
During normal operation the CLX will run through a timed cycle. A simplified cycle will
consist of the following sequences:
• Flushing – continuous sample flow
• Zeroing – no flow with cuvette full
• Purging – PURGE valve opens
• Adding Reagents – one pulse of the reagent pump
• Mixing with sample – sample flow pulses in
• Reading resulting sample – no flow with cuvette full
• Purging – PURGE valve opens to remove reacted sample
The cycle above does not describe all the actions and testing that occurs. The CPU
continuously diagnoses the entire system for correct operation and sample water flow. If
an error occurs, a message is posted to the message queue.
The reagent is added by a single pulse of the reagent solenoid. When the reagents require
replacement, the PRIME button is pushed to bring new reagents into the system. The
reagent solenoid is pulsed several times to draw fluid from the two reagent bottles and fill
the tubes with new reagent. A complete purge takes less than a minute.
The SERVICE button empties the cuvette, stops the flow of sample water, and clears any
errors. This provides a convenient way to replace reagents and the measurement cuvette. If
more extensive servicing is performed, all power to the CLX should be removed. If the
CLX is to be turned off, it is recommended that the instrument be placed in SERVICE
mode before removing power. This ensures that the cuvette is emptied and the flow is off.
CLX (9/07)
REV 3.1
Page 5
4.0
Installation and Commissioning
Prior to use for the first time, one of the reagents (the indicator) will have to be mixed.
Refer to section 10.2 Replacing or Installing the Reagents.
4.1 Mounting & Site Selection
The instrument is designed for wall mounting. If wall mounting is not practical, the
instrument can be mounted on any suitable level surface Choose a location that is easily
accessible for operation and service and ensure that the front display rests at eye level.
Consideration must be made the plumbing connections. The overall mounting dimensions
of the instrument are shown in Figure 3.
219.08mm
(8.625")
219.08mm
(8.625")
298.45mm
(11.750")
REAR VIEW
OF THE
INSTRUMENT
298.45mm
(11.750")
IT MAY BE NECESSARY TO REMOVE
THIS BULKHEAD NUT TO ACCESS THE
BOTTOM LEFT MOUNTING HOLE WITH
THE BRACKET IN THIS POSITION.
Figure 3: Overall Mounting Dimensions of the Instrument
It is critical that the instrument be mounted as close as possible to the sampling point to
ensure a quick response time (within 2-3 meters (6-10 ft) of the sampling point).
The provided mounting feet will need to be installed with the provided screws. These can
be rotated as shown above. Suggested mounting screws are up to M6 (¼”) .
CLX (9/07)
REV 3.1
Page 6
4.2 Plumbing
The recommended plumbing for the instrument is shown in Figure 4. The instrument is
designed to require very little head pressure to operate, but will need around 34.5kPa (5
PSI). The maximum pressure for proper operation should not exceed 276KPa (40 PSI).
The maximum allowable fluid temperature is 40°C (104°F).
FLOW
ADJUSTMENT
SCREW
PRESSURE
REGULATOR
CLX Assembly
CABINET VENT
DRAIN CONNECTION
ACCEPTS 12,7 MM
(1/2") I.D. TUBING
WATER INLET
ACCEPTS 6,35 MM
(1/4") O.D. TUBING
INTERMEDIATE TUBING
(SUPPLIED)
INLET STRAINER
(SUPPLIED)
NEW WATER INLET
ACCEPTS 6,35 MM
(1/4") O.D. TUBING
MAKE SURE
THAT THE
ARROW IS
POINTED TOWARD
THE CLX
Figure 4: Recommended Plumbing for the Instrument
The supplied T-strainer should be used to prevent clogging of the instrument. ¼” quick
connect fittings are supplied on the T strainer. If a pipe fitting change is required, this
change should be made at the T-strainer. Opaque tubing should be used if the tubing will
be exposed to sunlight, to prevent algae growth. Please note that the supplied connectors
are compatible with ¼” O.D. semi-ridged or ridged tubing.
The instrument is equipped with an internal cabinet drain (vent) to prevent damage in the
event of a tubing failure.
The drain tubing connects to a hose barb. The rated tubing size is ½” ID tubing. It is
recommended that opaque tubing be used to prevent algae growth. Keep this tubing as
short as possible. This drain must be kept open to the atmosphere.
CLX (9/07)
REV 3.1
Page 7
The fluid waste from drain connection of this instrument contains reagents diluted
with large quantities of sample water. HF scientific inc. recommends that operators
check with local authorities concerning proper disposal of waste fluids.
A ½ “ ID tubing can be placed over the cabinet vent to redirect accidental spills of
reagent to a suitable container.
4.3 Setting Flow Rate
The flow rate on the CLX was factory adjusted; however, installation variances may affect
the flow. The optimal flow rate through the CLX should be adjusted. The flow is adjusted
by turning the adjustment screw on the pressure regulator. Refer to figure 4. To assist in
this adjustment follow the procedure shown below:
1.
2.
3.
4.
5.
Press the SERVICE button.
Wait for the display to read HOLd, then press Mode/ Exit.
Display will show FLOW with the number 0. Press either the S or T button.
CLX will drain, and then pulse in water while a count is displayed on the screen.
The display will show one of three messages HI, LO or Good.
The flow test determines if the flow rate is suitable for proper operation. Adjust the
pressure regulator using a coin or a large flat blade screwdriver. Press either the S or T
button while in the FLOW routine to display a new flow rate. Please note that only ¼ turn
incremental adjustments should be made to the regulator on each attempt.
If the message is LO, turn the regulator control clockwise. If the message is HI, turn the
regulator counterclockwise. If the message is Good, no adjustment is required.
To return to normal operation, press the ↵ button.
4.4 Electrical Connections
All of the electrical connections to the instrument are made at the termination area which
is located on the left side of the instrument. Remove the high voltage cover by loosening
the captive screw. Refer to figure 5. The connections are labeled and are self-descriptive
(see Figure 5). Please follow all local and government recommendations and methods for
installation of electrical connections to and between the instrument and other peripheral
devices.
Plugs are inserted into the RS-485 and 4-20mA cable bulkheads when shipped, to ensure a
watertight seal. These plugs should be removed and discarded when cabling to either of
these connections.
The power cable bulkhead will accept cable diameters from 5.8mm (.230 in.) up to 10 mm
(.395 in.). All terminals are designed to accept wires in the range of 14-28 AWG. All
wires should be stripped to a length of 6 mm (¼”). A strain relief strap is provided to
reduce tension on the power terminals.
It is the user’s responsibility to assure that a watertight seal is maintained after the CLX
has been wired for operation. If any of the bulkheads are not tightened properly around a
cable or plug, the ratings of the instrument will be jeopardized and there is a possibility of
creating a shock hazard.
CLX (9/07)
REV 3.1
Page 8
Note: Only qualified electricians should be allowed to perform the installation of the
instrument as it involves a line voltage that could endanger life.
CAPTIVE SCREW
HIGH VOLTAGE COVER
POWER CABLE
STRAIN RELEIF
ANCHOR
POWER
TERMINAL
BLOCK
POWER
BULKHEAD
ALARM #2
TERMINAL
BLOCK
ALARMS
BULKHEAD
ALARM #1
TERMINAL
BLOCK
RS-485
TERMINAL
BLOCK
COMMUNICATION
BULKHEAD
4-20mA
TERMINAL
BLOCK
Figure 5: Electrical Connections for the Instrument
4.4.1 Power
The instrument is equipped with 100-240 VAC, 47-63 Hz power supplies requiring
150VA; please verify that the line voltage falls within these specifications. It is
recommended that a circuit breaker be placed prior to the power connection to
allow for service. For safety it is recommended that the connection be less than 2
meters (six feet) from the instrument. While making connections, refer to Figure 5.
The CLX is intended for cord connection with a three wire non-locking grounded
power cord. A power cord can be purchased separately from the factory (Catalog
No. 20779S). The CLX is not supplied with a power cord. If the CLX is to be
used in the U.S. or Canada the power cord must be UL Listed & CSA Certified.
Please consult all local electrical codes for proper connection.
CLX (9/07)
REV 3.1
Page 9
The connection block is marked N for Neutral and L for line the third symbol
indicates a secure earth ground. The green removable terminal block is suitable for
wire gauges 18 to 12 AWG.
4.4.2 RS-485
The RS-485 half-duplex (2-wire) digital interface operates with differential levels
that are not susceptible to electrical interferences. This is why cable lengths up to
3000 ft can be implemented. The last device on each bus may require terminating
with a 120 ohm resistor to eliminate signal reflection on the line. Do not run RS485 cables in the same conduit as power. Set-up of the RS-485 is covered in 7.3
Configuring the RS–485 Port
To prevent damage to the instrument, ensure that power is disconnected prior to
making connections. For ease of connecting, remove the plug-in terminal block.
Connections are labeled beside this termination on the PC board.
The recommended cable is 22 AWG shielded twisted pair. The grey terminal block
is removable to assist in making connections.
4.4.3 Relays
The Alarm 1 and Alarm 2 relays are mechanical relays rated at 240 VAC 2A.
Please note that the relays are labeled NO (Normally Open), NC (Normally
Closed) and C (Common). As these alarms are configured fail-safe, the normal
condition is with power applied to the CLX and in a non-alarm condition.
Operation of these alarms is covered in section 7.4 Configuring the Alarms.
The lever operated terminal blocks are rated for wire gauges 28-14.
4.4.4 4-20 mA
The 4-20 mA output is driven by a 15 VDC power source and can drive recorder
loads up to 600 ohms. Transformer isolation is provided on the CLX. Do not run 420 mA cables in the same conduit as power. Operation of this output is covered in
section 7.1 Setting the 4-20 mA.
To prevent damage to the instrument, and for general safety ensure that power is
disconnected to the CLX prior to making any connections. For ease of connecting,
remove the plug in terminal block. Polarities of the connections are labeled beside
this termination on the PC board.
The recommended cable is 22 AWG shielded twisted pair. To prevent ground
loops, connect the shield at either the CLX or at its destination, but not both. The
grey terminal block is removable to assist in making connections.
4.4.5 RS-485/4-20 mA cable Ferrite
To meet IEC requirements for RF radiated immunity a clamp-on type ferrite is
supplied. It should be placed on the RS-485 or 4-20 mA cable outside, but as close
as possible to the CLX. If both outputs will be used, an additional ferrite will be
needed and can be ordered from HF scientific inc.
CLX (9/07)
REV 3.1
Page 10
5.0
Operation
The CLX Online Chlorine Analyzer allows for the measurement of the chlorine of process
water on-line. The chlorine value of the process water is usually reported in milligrams
per Liter (mg/L), these units are equivalent to Parts Per Million (PPM).
Readings above 10.00 mg/L are outside the range of this instrument. Readings above
6mg/L may not be within the stated accuracy.
5.1 Routine Measurement
First, ensure that all plumbing and electrical connections are complete before continuing.
The following steps describe how to measure the value of chlorine of a sample using this
instrument:
1. Apply power to the instrument and allow the unit to warm up (typically 45 minutes to
one hour on initial commissioning).
2. When a continuous process stream is flowing through the instrument, the instrument
will display the measured chlorine level of the sample by displaying it on the LCD
screen. In addition, the equivalent signal is provided on the analog (4-20 mA) output,
or the digital output, depending on the options selected.
During normal operation, the instrument will have the arrow beside AUTO highlighted
with the current scale displayed on the lower row of the display and the measured reading
on the upper row of the display (see illustration below).
The screen depicted below indicates that the system has just been started or just entered
AUTO mode from Service mode and no readings have been taken yet. Please note that
calibrations will not be allowed until a reading is posted.
CLX (9/07)
REV 3.1
Page 11
5.2 Security Access Feature
The instrument is equipped with a security access code feature that can be activated in the
configuration mode. If the security feature is enabled, the screen shown in the illustration
below will appear when the MODE/EXIT button is pressed.
The security code (333) must be entered to gain access to CAL or CONFIG menus.
Notice that the first number in the code is flashing. The flashing indicates that this is the
number to be changed. Use the S or T arrows to select the first of the three numbers in
the code and then press the button to accept the first number of the code. Now enter the
second number in the code. Proceed as with the first number followed by . Then repeat
the process for the third number in the access code, and finish with the button.
If the valid access code has been selected, the instrument will be directed to the calibration
mode. If the wrong access code is selected, the instrument will return to the AUTO mode.
Refer to section 7.5 Enabling the Security Access for more information.
CLX (9/07)
REV 3.1
Page 12
6.0
Instrument Calibration
The instrument was tested prior to leaving the factory. Since it operates from a predetermined calibration curve, no calibration is required.
Calibration can easily be performed if required. The method is by comparison against
another instrument, such as a laboratory or hand held photometer (such as HF scientific’s
Chlorine Pocket Photometer).
There are two points of calibration. The slope or gain and the zero (offset). To perform the
zero the instrument must be plumbed to a sample of known chlorine free water such as deionized water for a zero adjustment.
6.1 Slope (gain) Calibration Procedure
It is important that the chlorine level be quite stable to use this method. The comparison
will be made against a trusted measurement such as a chlorine photometer,
spectrophotometer, or an amperometric titration.
1. Obtain a grab sample of the flow.
2. Measure the value of the sample with one of the methods shown above.
3. On the CLX, press the MODE/EXIT button once. The screen is shown below.
4. Press to enter the calibration adjustment.
5. The screen will show the current reading on the CLX. Using the S & T buttons
adjust the reading to agree with the laboratory method or portable photometer.
6. Press to accept the calibration adjustment and return to AUTO measurement
mode.
Notes: There is a limit to the size of the change that can be made to a current reading. The
upper limit is the current reading times 1.5. The lower limit is the current reading
divided by 1.5.
Ensure a reading is posted to the display before calibrating to avoid a nOnE error.
CLX (9/07)
REV 3.1
Page 13
6.2 Zero (Offset) Calibration Procedure
Generally this calibration is only required if readings are expected to be below 1 mg/L or
if it is required by a regulatory authority. To perform this calibration, the water supply to
the CLX must be changed to chlorine free water such as de-ionized water. This chlorine
free water must be run through the instrument for at least 5 minutes prior to using the
following procedure.
1. On the CLX, press the MODE/EXIT button once. The screen is shown below.
2. Press either the S or Tbuttons to get the following screen.
3. Press to enter the zero calibration screen.
4. The screen will show the current reading on the CLX. Since there is no
chlorine, this is a zero offset.
5. Press to Zero the instrument and return to AUTO measurement mode.
Note: There is a limit of ±0.20 mg/L total adjustment available. A ZERO Cal. greater than
this will cause a CAL warning and no calibration will have occurred. Enter
SERVICE mode to clear this error.
6.3 Restore Factory Settings
If the CLX displays a CAL error or the calibration was incorrectly performed, it may be
desired to restore the factory calibration. All factory defaults including factory
configurations can be reset by holding down the S button and then pressing and releasing
the button then releasing the S button.
CLX (9/07)
REV 3.1
Page 14
7.0
Instrument Configuration (CONFIG mode)
The instrument has been designed to provide the ability to customize the instrument
according to needs at any time during normal operation. This mode has been split into
sub-menus to facilitate instrument configuration. This section describes how to use each
of the sub-menus to configure the instrument. While in the configuration mode, the
instrument has a time-out feature that automatically returns the system operation to the
AUTO mode after a fifteen (15) minute period of no button pushes.
Enter the CONFIG mode of the instrument by pressing the MODE/EXIT button until the
arrow beside CONFIG is illuminated, then press the ↵ button.
Note: To exit the CONFIG mode, press the MODE/EXIT button.
7.1 Setting the 4-20 mA Output
The first configuration selection is 4-20 for the 4-20 mA output. Select the either On or
OFF using the S and T buttons. Once the desired output has been set, press the button
to accept it. The next prompts will depend on the output selected.
If the 4-20 mA output was turned On, prompts to set the 4mA(4MA) and 20mA(20MA)
chlorine limits will be displayed. There will also be a menu to adjust the error level
(ERLV). The first prompt will be the chlorine limit assigned to the 4 mA output level:
Select the chlorine level to assign to the 4MA using the S and T buttons.
Once the desired level has been set, press the button to accept it.
Note:
The 4MA can be set higher than 20MA level to invert the output current if required.
This may be required to control a dosing pump.
CLX (9/07)
REV 3.1
Page 15
The next prompt will be the chlorine level assigned to the 20MA. Select the chlorine level
using the S and T buttons. Once the desired level has been set, press the button to
accept it.
7.2 Configuring the Error Level
In case of an error in the CLX, the 4-20 mA reading can be used to indicate a problem by
sending the current to either 4.00 mA, 2.00 mA or 0 mA. If this type of error indication is
not wanted, it can be set to OFF. OFF should always be used if the CLX is used to control
dosing.
7.3 Configuring the RS–485 Port
CLX (9/07)
Page 16
REV 3.1
The instrument is equipped with an RS 485 port which operates in Simple bus, a
proprietary communication (for HF Online) or Modbus. Prompts will appear for setting
the baud rate, the address and the Modbus transmission mode (RTU or ASCII).
Select the correct baud rate (1200, 2400, 4800, 9600, or 19200) for operation of the I/O
port by pressing the S or T buttons to change the displayed baud rate.
Press the button to continue on and select the desired instrument address using the S or
T buttons. Once the selection is satisfactory, press the button.
To use the Modbus mode, select ASCII or RTU. Refer to the Modbus Manual available
from HF scientific inc. or online at www.hfscientific.com.
7.4 Configuring the Alarms
Two relays are provided that are designed to operate as two independent programmable
alarms or as a system problem alarm. Please note that changes to alarms will not be
recognized until the start of the next cycle. Two settings must be selected to fully program
each alarm:
1. The alarm function (HI, LO, OFF or Error)
2. The alarm set point (level at which the alarm activates)
These items are described below:
Alarm Function: The alarms can either be turned OFF or selected to operate in one of
three different manners:
1. HI alarm: the relay changes state when the measured chlorine level is higher than the
programmed alarm level (set point).
2. LO alarm: the relay changes state when the measured chlorine level is lower than the
CLX (9/07)
REV 3.1
Page 17
programmed alarm level (set point).
3. Error alarm: If there is a system fault or problem the alarm will change states.
Alarm Set Point: The level at which an alarm activates is called the alarm set point. On
the instrument, the alarm set point is designated as “S/P”. The set point is adjustable to
any valid chlorine level over the range of the instrument in steps of 0.01 mg/L. This
setting is not available if the Error function is chosen
7.4.1 Alarm 1
Alarm 1 Function: The ALM1 is displayed and the display indicates the current
function of alarm 1 (HI, LO, OFF, or Error). Use the S or T buttons to cycle
through and select the desired function. Press the button to accept the selection.
If the alarm was turned OFF, a prompt will appear to set up alarm 2 (go to section
7.4.2).
Alarm 1 Set Point: This prompt is used to select the set point for this alarm; this
is indicated by “S/P” shown on the lower row of the display. Select the desired
alarm level by using the S and T buttons. Once the desired set point has been
set, press the button to accept it.
7.4.2 Alarm 2
Repeat the procedure listed in section 7.4.1 to set up the parameters for alarm 2. If
a selection was made to turn the alarm OFF, the next selection for the speed of
response RESP is shown.
Note: Due to the cyclic nature of the CLX, relay chatter is not an issue. There is no need for
alarm delays or hysteresis.
7.5 Enabling the Security Access
The instrument is equipped with a security access. If this option is turned on, the user is
required to input the access code into the instrument to get to any mode other than AUTO.
The only code is 333. This code may not be changed. See section 5.2 for more
information on this security feature. The security key icon will be visible and flashing on
the display whenever the access option is selected using the S or T buttons. (On or
OFF).
CLX (9/07)
REV 3.1
Page 18
7.6 Extended Settings
The last few settings are grouped together to prevent them from being adjusted by
accident. To gain access to the extended settings, select On using the S or T buttons and
press the button.
7.7 Speed of Response
The speed of response for both displayed and output values of mg/L can be adjusted in
this menu. Although the default setting is 10, sixty-one (0-60) response speeds are
available. The displayed number is a response time, in minutes. Cycle Times do not affect
the response time. Select the desired speed of response using the S and T buttons. Press
the button to accept it.
To avoid reading air and other anomalies, select the slowest speed (60). Select the fastest
response (0) where monitoring of rapid changes is needed.
7.8 LCD Backlight Brightness
The LCD backlight brightness may need to be adjusted. This is of particular interest if
multiple instruments are located in the same area and it is desired for the entire group to
have the same appearance. Ten levels are available. The default brightness is 8.
Change the brightness by pressing the S or T button. When the desired brightness has
been selected, press the button.
CLX (9/07)
REV 3.1
Page 19
7.9 RS-485 Parameters
These menus will only appear if the RS-485 is enabled (see 7.3). The default is 8 Bit, no
(nOnE) Parity, 1 Stop Bit. Make selections using the S and T buttons then press the button to move to the next menu.
7.10 Cycle Time
The cycle time can be changed using this menu. Please note that changing this menu
will directly affect the volume of reagent that will be consumed. The default is set to
150 seconds (2 ½ minutes). Using this setting the reagents will last 30 days. Make
selections using the S and T buttons then press the ↵ button to move to exit to AUTO
mode. Allowable setting is from 90 to 600 seconds (10 minutes).
7.11 Saving Configuration Settings
If extended settings are set to OFF, pressing the button will save all settings and the
CLX will automatically return to the normal AUTO mode of the instrument.
If extended settings are set to On, after the last menu of the extended settings, pressing the
button will save all settings and the CLX will automatically return to the normal AUTO
mode of the instrument.
The CONFIG menu may be used at any time to reset or change any of the parameters.
The CONFIG menu may be exited at any point in the menu by using the MODE/EXIT
key. Any features that have been modified will be saved.
CLX (9/07)
REV 3.1
Page 20
8.0
Additional Features and Options
8.1 Backlit LCD
The backlit LCD allows for easier readability of the LCD display in low light or no light
conditions. The backlight is intended for continuous operation. The brightness is
adjustable from a menu in the CONFIG mode.
8.2 RS-485 Outputs
The CLX has the capability to operate in three different RS-485 modes. Included is a
mode for interfacing into the HF Online software package (section 8.2.1 below), a simple
communication mode and Modbus communications. All modes will automatically
configure and do not require any changes or selections
8.2.1 HF Online (HF catalog # 19783)
The CLX can operate as a small SCADA system with an optional PC software
package, called HF ONLINE. This system allows for an interface with a
combination of Micro TOL’s or CLX’s, totaling up to 255 units, for the purpose of
data logging. This system will interface directly with common database and
spreadsheet software.
8.2.2 Simple Communication
The CLX can provide basic communications over simple programs such as the
Hilgraeve HyperTerminal that is included with most Microsoft Windows
packages. The user could also use Visual Basic or other programs. The default
communication parameters are 8 bits, no parity and 1 stop bit. These can be
changed in the Extended CONFIG menus 7.9 RS-485 Parameters.
The master computer will send out:
•
Byte #1 the attention character “:” in ASCII or 3A Hex
•
Byte #2 the address of the CLX being queried
•
Byte #3 & 4 CR LF or 0D 0A in hex
The CLX will respond with:
•
The same attention character “:” in ASCII or 3A Hex
•
The address of the CLX
•
The Reading
•
The Unit (mg/L)
A sample communication would look like this:
CLX (9/07)
REV 3.1
(Master computer requesting a report from address #1)
:1 CRLF
(CLX set to address #1 Response)
:001 0.0249mg/L
Page 21
8.2.3 Modbus Communication
Modbus protocol communication manual is available HF #24569. This manual is available
free online at www.hfscientific.com.
8.3 Remote Panel Meter (Catalog # 19609)
The remote panel meter allows for remote indication of the mg/L reading using the 4-20
mA loop of the CLX. No external power is required, as the meter is run off of the 4-20
mA source of the CLX.
8.4 Desiccant Cartridge Option (Catalog #09944)
An optional desiccant cartridge kit can be purchased for use in applications where
condensation on the glass cuvette may compromise accuracy. The desiccant changes color
from blue to pink when expended. Instructions for replacement are included with the kit.
CLX (9/07)
REV 3.1
Page 22
9.0
Troubleshooting
9.1 CLX Fault Detection
The CLX performs continuous diagnostic monitoring. In the CLX, there are 4 severity
levels of fault detection. Level 4, 3 & 2 will allow normal operation, but warn of the
problem. Level 1 is an instrument failure and the instrument will not operate. Any faults
are displayed in a queue form in the bottom row of the LCD.
A level 4 fault is simply a screen indication that one of the alarm levels has been
activated. This fault level will not affect the 4-20 mA and will only affect the alarm
activated. The sample back light blinks at a rate of once every 4 seconds.
A level 3 fault indicates a failure or a problem that usually can be corrected by the
operator. Refer to the chart below. If any of these errors occur, the instrument will still
display readings and probably will operate correctly. These faults will self-clear when the
problem is corrected. If any of these faults occur, they will affect the 4-20mA and any
alarm set for fault detection (Error). The sample back light blinks at a rate of once every
2.5 seconds.
Message
MA
Description of Fault
4-20 mA enabled & loop open
Corrective Action
Check 4-20 mA wiring or turn off 420mA if not used
CAL
Calibration invalid –not accepted
Recalibrate if needed
WATR
No water flowing
Check water flow
FAST
Intake water flow too fast
Set flow rate (see section 4.2.1)
SLOW
Sample cuvette filling too slowly
Check water flow
PURG
Sample cuvette has slow purge
Check drain lines
NPRG
Sample cuvette not purging
Check drain lines
H2O
CLX cannot determine water level
Check water flow
ISOL
Problem with intake solenoid
Check wiring, check for clogged solenoid
PSOL
Problem with purge solenoid
Check wiring, check for clogged solenoid
RGNT
Problem with reagent
Check reagent and lines
GLAS
Dirty cuvette
Replace or clean cuvette
WCAL
Water Level Calibration Invalid
Clear fault (see section 9.2)
REPL
Reagents Expired
Replace reagents and press PRIME
A level 2 fault indicates a severe problem that will usually require technical assistance
from HF scientific customer service (see section 9.4). The queued display will show
POST. If this fault occurs it will affect the 4-20mA and any alarm set for fault detection
(Error). The sample back light blinks at rate of once every 1 second.
A level 1 fault is a system fault. This is NOT a problem that the operator can correct, and
the unit must be returned to the factory for service (see section 9.4). These failures consist
of failures in the CPU, A/D, EEPROM or other devices internal to the instrument. The
queued display will show FAIL, the upper display is a five digit code. If this fault occurs,
it will affect the 4-20mA and any alarm set for fault detection (Error). The instrument will
not operate with this fault. The sample back light blinks at rate of once every 0.4 seconds.
If any fault conditions occur, the message indicating the fault will be shown on the lower
row of the display.
CLX (9/07)
REV 3.1
Page 23
9.2 Clearing Faults
Every time SERVICE mode is exited, all faults are cleared. If the original fault or a new
fault occurs, it will be posted.
9.3 Diagnostic Chart
Symptom
Cause
Lower display shows MA
4-20 mA loop open
Lower display shows FAIL
Readings are erratic
Major system fault
(1) Bubbles in solution
(2) Debris in flow
Readings are lower than expected
(1) Condensate or leaky
measurement cuvette
(2) Measurement cuvette
dirty
(3) Reagents bad or expired
Sample Over-Range
Upper display flashes
Upper display shows nOnE while
attempting to calibrate
No current reading displayed
Cure
Check wiring. See sections
4.4.4 and 7.2
Refer to section 9.1 & 9.2
(1) See above
(2) Install T strainer at inlet
(1) Install desiccant cartridge
kit
(2) Replace or clean cuvette
(3) Replace reagents
Check sample. Sample may be
too high to read.
Wait for CLX to post a reading
9.4 Technical and Customer Assistance
If for any reason assistance is needed regarding this instrument please do not hesitate to
contact either the HF scientific, inc. Technical Service Department or the HF scientific,
inc. Customer Service Department:
HF scientific, inc.
3170 Metro Parkway
Fort Myers, Florida 33916-7597
Phone: (239) 337-2116
Fax: (239) 332-7643
Email: [email protected]
www.hfscientific.com
CLX (9/07)
REV 3.1
Page 24
10.0 Routine Maintenance
10.1 Maintenance Schedule
The recommended schedule is shown below. It is important to replace the reagents on a
monthly basis to get reliable accurate readings from the CLX.
The CLX is shipped with one CLX Tubing/Cuvette kit, HF part # 09950. The kit consists
of the following:
Part
Qty
8
“B” tubings
2
“C” tubings
2
“D” tubings
1
Cuvette
CHECK VALVE
REAGENT BOTTLE CAP
"B"
"C"
"D"
The drawing above shows one complete reagent tubing set. Two complete sets are used in
the CLX; one for the buffer and one for the indicator.
The supplied kit is intended to last for one year. Additional kits can be ordered from your
local HF scientific inc. distributor or representative. It is recommended to keep one kit on
hand at all times.
Generally, all “B” tubes should be replaced on a three month basis. The “C” and “D”
tubes and the cuvette should be replaced annually. If the CLX is operated in an ambient
temperature above 27 ° C (81° F), the “B” tubes should be changed every two months.
Every Month
1. The reagent required for operating this instrument must be changed on a monthly
basis (with a 2.5 minute cycle time).
2. The external strainer (if used) should be checked and cleaned if necessary
3. The glass cuvette should be inspected. Check for excessive debris on the inside
surface of the glass. It is suggested to keep a spare cuvette to replace when required.
The old cuvette may be cleaned, if possible, for future replacement.
CLX (9/07)
REV 3.1
Page 25
Flushing the System
It is recommended that the tubings replacements be timed with reagent replacement. Press
the SERVICE button to stop the water flow. Remove old reagents and discard. Place the
inlet tubings in a small container of clean water. Press SERVICE to return to operation
mode, press PRIME and then to flush the system with water. Remove the inlet tubings
from the water Press PRIME and then to remove most of the water.
Note: After a PRIME the CLX will perform a water calibration (WCAL). It will take a few
minutes to complete this procedure.
3 Month Tubing Replacement
The “B” tubings need replacement more often due to the fact that they are subject to wear
from the reagent “Pump”. The check valves should not need to be replaced and should be
saved. Please note that the check valves are directional and that the “IN” side is smaller in
diameter (see drawing).
Steps:
1 Flush the system as described above to reduce personal contact with the reagents.
2 Press SERVICE to stop the flow of sample water and drain the cuvette.
3 Remove and retain the thumb screw and washer on top of the pump; pull the pump
hammer up and out of the way of the “B” tubing. Do not remove the hammer
completely.
4 Work on one reagent side at a time. Disconnect the “C” tubing between the reagent
cap and the check valve.
5 Remove the check valve from its seat by pulling it down and out.
6 Replace the “B” (black) tubing between optics inlet and the check valve OUTLET.
Discard the old “B” tubing.
7 Replace the check valve into its seat.
8 Reconnect the “C” tube (from the reagent bottle caps) to the check valve INLET.
9 Repeat steps 4-8 for the other reagent.
10 Check the drawing on the following page to ensure correct installation.
11 Return to operation as described.
Annual Tubing Replacement
Once a year, complete the 3 Month Tubing Replacement as above, and then follow the
steps below:
1. Replace the two “C” tubings (from the check valves to the reagent bottles).
2. Replace the two “D” tubings (Reagent Bottle intake tubing).
Note: “D” tubing is notched on one side. This notched side should face the bottom of bottle.
3. Turn the knurled top on the optics system counterclockwise (as viewed for the top)
until the cuvette just “pops” out, but do not remove the top.
4. When the cuvette “pops” out, move the retaining o-ring & remove the cuvette. You
may need a stiff wire such as a bent paper clip to grasp the cuvette. Retain this cuvette
for future use if it can be cleaned.
5. Install the new cuvette by pushing it firmly in place and turning the knurled top
clockwise until the cuvette is held securely.
6. Check the drawing on the following page to ensure correct installation.
7. Return to operation as described.
CLX (9/07)
REV 3.1
Page 26
Return to Normal Operation
Press the SERVICE button to return sample flow to the system. Check for leaks. If a leak
occurs press SERVICE again, repair leak and try again. Once the system is operating
correctly, return or replace reagents and press PRIME and then one time to restart
reagent flow. The system will automatically return to normal operation.
Notes: Tubes may darken due to contact with the reagent. This condition does not affect the
performance of these parts.
KNURLED TOP
REAGENT
BOTTLE
CAP
OPTICS SYSTEM
TUBING "C"
THUMB
SCREW
TUBING "D"
CUVETTE
RETAINING
O-RING
TUBING "B"
PUMP
HAMMER
INLET
CHECK
VALVES
REAGENT
PUMP
ASSY
CLX (9/07)
REV 3.1
FLOWy
INLET
OUTLET
CHECK VALVE DETAIL
Page 27
10.2 Replacing or Installing the Reagents
Reagent kits are available from HF scientific Inc. Catalog # 09947 for Free Chlorine and
Catalog # 009948 for Total Chlorine. There are two reagents required, the buffer and the
indicator. The buffer comes premixed and ready to use.
Note: The indicator solution must be prepared. Avoid preparation of this reagent in direct
sunlight as this may lower the reagent life.
Add the contents of one bottle of DPD Powder to a bottle of Free/Total Indicator Solution.
Recap the bottle & shake for 2-3 minutes until the powder is dissolved completely. Check
the bottle to ensure that the powder is completely dissolved prior to use.
Please note that life of the mixed indicator solution is 1 month. Old standards should be
discarded.
To replace the reagents, press the SERVICE button; this will empty the cuvette and stop
any flow of water. Remove the cap on both bottles replace with the cap supplied in the
CLX. Be sure to replace the reagents in the correct location as labeled. The suction tube
for both reagents will reach the bottom of the bottles.
To complete the replacement procedure, press the PRIME button. This will draw enough
of each reagent to completely prime the tubes and replace any old solution. The system
will automatically return to normal operation after it has primed.
Note: It may take about ½ hour after priming before accurate readings are displayed.
Use caution when changing the reagents as the reagents are quite strong.
These reagents will stain clothing and anything that they contact. After changing
the reagents, operators should wash their hands.
10.3 Cuvette Change
As recommended in section 10.1 Maintenance Schedule the cuvette should be inspected at
monthly intervals and replaced when needed.
To replace the cuvette, press the SERVICE key. When HOLD shows on the screen, the
system is ready.
Warning: The cuvette is spring loaded to allow for removal. Leave the cuvette retainer (Oring) in place while loosening top.
Loosen the knurled top until the cuvette pops out. Pull the cuvette retainer down to allow
for cuvette removal. Remove and replace or clean cuvette. Reverse the procedure to
replace. To avoid damage do not over tighten the knurled top. Press the SERVICE key to
return the system to normal sample flow and check for leaks.
CLX (9/07)
REV 3.1
Page 28
10.4 Instrument Storage
If the CLX is relocated or will be inactive for more than 48 hours, remove the reagents.
Flush the reagent system as describe in 10.1 Maintenance Schedule. Place the instrument
in Service mode to drain the system then remove power by disconnecting the mains power
plug. It is usually a good idea to disconnect or shut off the source water.
10.5 Cleaning the CLX
Flush the system as mentioned in section 10.1Maintenance Schedule. When the flushing is
finished, press the SEVICE button wait until the display reads HOLd.
As a mater of safety, always disconnect any power source to the CLX prior to attempting
any cleaning. It is recommended that the source water is also shut off.
Isopropyl alcohol (rubbing alcohol) on a soft cotton cloth works very well in removing
reagent stains from plastic parts, the key pad and the display. Use care when cleaning
around electrical components. Do not use any harsh cleaning agents as these may cause
damage to the instrument components.
Ensure that the system is dry prior to applying power.
CLX (9/07)
REV 3.1
Page 29
11.0 Accessories and Replacement Parts List
The items shown below are recommended accessories and replacement parts.
Accessory
Catalog Number
Reagent Kit – Free Chlorine 30 day supply
09947
Reagent Kit – Total Chlorine 30 day supply
Operating Manual CLX
09948
25016
Tubing/Cuvette Kit
09950
Replacement Cuvette
25018S
Check Valve Set
25017S
To order any accessory or replacement part, please contact the HF scientific, inc.
Customer Service Department. If for any reason technical assistance is needed regarding
this instrument, please do not hesitate to contact the HF Technical Services Department.
HF scientific, inc.
3170 Metro Parkway
Fort Myers, Florida 33916-7597
Phone: (239) 337-2116
Fax: (239) 332-7643
Email: [email protected]
www.hfscientific.com
CLX (9/07)
REV 3.1
Page 30
12.0 Warranty
HF scientific, inc., as vendor, warrants to the original purchaser of this instrument that it
will be free of defects in material and workmanship, in normal use and service, for a
period of one year from date of delivery to the original purchaser. HF scientific, inc.’s,
obligation under this warranty is limited to replacing, at its factory, the instrument or any
part thereof. Parts, which by their nature are normally required to be replaced
periodically, consistent with normal maintenance, specifically reagent, desiccant, sensors,
electrodes, tubing and fuses are excluded. Also excluded are accessories and supply type
items.
Original purchaser is responsible for return of the instruments, or parts thereof, to HF
scientific, inc.’s factory. This includes all freight charges incurred in shipping to and from
HF scientific, inc.’s factory.
HF scientific, inc. is not responsible for damage to the instrument, or parts thereof,
resulting from misuse, environmental corrosion, negligence or accident, or defects
resulting from repairs, alterations or installation made by any person or company not
authorized by HF scientific, inc.
HF scientific, inc. assumes no liability for consequential damage of any kind, and the
original purchaser, by placement of any order for the instrument, or parts thereof, shall be
deemed liable for any and all damages incurred by the use or misuse of the instruments, or
parts thereof, by the purchaser, its employees, or others, following receipt thereof.
Carefully inspect this product for shipping damage, if damaged, immediately notify the
shipping company and arrange an on-site inspection. HF scientific, inc. cannot be
responsible for damage in shipment and cannot assist with claims without an on-site
inspection of the damage.
This warranty is given expressly and in lieu of all other warranties, expressed or implied.
Purchaser agrees that there is no warranty on merchantability and that there are no other
warranties, expressed or implied. No agent is authorized to assume for HF scientific, inc.
any liability except as set forth above.
HF scientific, inc.
3170 Metro Parkway
Fort Myers, Florida 33916-7597
Phone: (239) 337-2116
Fax: (239) 332-7643
Email: [email protected]
Website:www.hfscientific.com
CLX (9/07)
REV 3.1
Page 31

advertisement

Key Features

  • Measures free chlorine and chloramines
  • Backlit LCD display
  • 4-20 mA output
  • Configurable alarms
  • Easy to install and use
  • Requires minimal maintenance

Related manuals

Frequently Answers and Questions

How often should I calibrate the Watts CLX Online Chlorine?
The Watts CLX Online Chlorine should be calibrated every 3 to 6 months, or more often if necessary.
What is the warranty on the Watts CLX Online Chlorine?
The Watts CLX Online Chlorine comes with a one-year warranty.
Can I use the Watts CLX Online Chlorine to measure chlorine in saltwater pools?
Yes, the Watts CLX Online Chlorine can be used to measure chlorine in saltwater pools.

advertisement