Jaycar QM1571 Operating Instructions Manual

Add to My manuals
24 Pages

advertisement

Jaycar QM1571 Operating Instructions Manual | Manualzz

QM1571 OPERATING INSTRUCTIONS

TRUE RMS AUTORANGING DIGITAL MULTIMETER www.procontechnology.com.au

Phone: (03) 98306288

SAFETY WARNINGS

The following safety information must be observed to insure maximum personal safety during the operation at this meter:

♦ Measurements beyond the maximum selected range must not be attempted.

♦ Extreme care must be taken when measuring above

30 VAC or 60VDC, especially on live bus-bars.

♦ To measure voltage, the instrument must not be switched to a current or resistance range, or to the diode check or buzzer position.

♦ Circuits must be de-energised and isolated before carrying out resistance tests.

♦ The rotary function switch must only be turned after removing test connections.

♦ All external voltages must be disconnected from the instrument before removing the battery.

Test leads and probes must be in good order, clean, and with no broken or cracked insulation.

UK Safety Authorities recommend the use of fused test leads when measuring high energy systems.

♦ Replacement fuses must be of the correct type and rating.

♦ The instrument must not be used if any part of it is damaged.

www.procontechnology.com.au

Phone: (03) 98306288

Warnings and precautions must be read and understood before the instrument is used. They must always be observed during the operation of this instrument.

Symbols used on this instrument are:

Caution: refer to accompanying notes

This symbol indicates that the operator must refer to an explanation in the

Operating Instructions to avoid personal injury or damage to the meter.

Caution: risk of electric shock

This WARNING symbol indicates a potentially hazardous situation, which if not avoided, could result in death or serious injury.

This CAUTION symbol indicates a potentially hazardous situation, which if not avoided, may result in damage to the product.

www.procontechnology.com.au

Phone: (03) 98306288

This symbol advises the user that the terminal(s) so marked must not be connected to a circuit which the voltage with respect to earth or ground exceeds

(in this case) 1000 VAC or VDC.

Equipment protected throughout by

Double Insulation (Class III).

Equipment complies with current EU directives.

SYMBOLS AND ANNUNCIATORS

Continuity

Low Battery

Wireless transmit mode

Diode test

HOLD Data Hold

AUTO

AC

Auto Ranging

Alternating Current or Voltage

www.procontechnology.com.au

Phone: (03) 98306288

DC

V

A, mA,

µ

A

Direct Current or Voltage

Volts

Amps, milli-Amps, micro-Amps

OPERATION

To turn on the meter rotate the function switch from the

OFF position to any measurement position.

For best battery life, ALWAYS turn the function switch to the OFF position when the meter is not in use. This meter has an Auto OFF capability that automatically shuts the meter OFF after 30 minutes.

NOTE: On some low AC and DC voltage ranges, with

the test leads not connected, the display may show a random, changing reading. This is normal and is caused by the high-input impedance and the input sensitivity.

The reading will stabilize when connected to a circuit.

MODE button

Selects AC or DC measurement when in Voltage or

Current modes, also selects resistance, diode, continuity or capacitance modes and °C or °F display modes.

www.procontechnology.com.au

Phone: (03) 98306288

HOLD/BACKLIGHT button

The HOLD button holds the current display reading until momentarily pressed again. Press and hold this button for more than 2 seconds to turn the backlight on. Press again for more than 2 seconds to turn off the backlight.

RANGE button

When the function switch is operated, the meter automatically starts operating in Auto Ranging mode.

This selects the best range for the measurement being made and is generally the best mode for most manual measurements. For data logging measurements or when required, the range may be manually selected, as follows:

1. Press the RANGE button. The "AUTO" indicator on the display will turn off.

2. Press the RANGE button to step through the available ranges until the required range is selected.

3. Press and hold the RANGE button for more than 2 seconds to revert back to "AUTO" operation.

www.procontechnology.com.au

Phone: (03) 98306288

REL button

The REL button performs a relative measurement.

Except for Hz, %, Diode and Continuity, all other functions will operate in the relative mode.

Hz% button

Hz% is the Frequency/Duty Cycle select button. In the frequency measurement mode (function switch set to

Hz%), pressing the button can select frequency or duty cycle measurements; in AC/DC Voltage or Current modes, pressing the key will switch to Frequency/Duty

Cycle measurement mode and then back to voltage or current mode.

DC/AC VOLTAGE MEASUREMENT

1) Insert the black test lead into the negative COM jack and the red test lead into the positive V jack.

2) Set the function switch to the Voltage V position.

3) Use the MODE button to select AC or DC Voltage.

4) Connect the test leads to two points in the circuit to measure the voltage difference.

5) Read the voltage measurement on the display.

www.procontechnology.com.au

Phone: (03) 98306288

DC CURRENT MEASUREMENT

1) Insert the black test lead into the negative COM jack.

2) For current measurements up to 4000

µ

A DC, set the function switch to the

µ

A position and insert the red

test lead banana plug into the

µ

A jack.

3) For current measurements up to 400mA DC, set the function switch to the mA position and insert the red test lead banana plug into the mA jack.

4) For current measurements up to 10A DC, set the function switch to the 10A position and insert the red test lead banana plug into the 10A jack.

5) Use the MODE button to select AC or DC Current.

"DC" must appear on the display.

6) Remove power from the circuit under test, then open the point in the circuit where you wish to measure current.

7) Touch the black test probe tip to the negative side of the circuit. Touch the red test probe tip to the positive side of the circuit.

8) Apply power to the circuit.

9) Read the DC current measurement on the display.

www.procontechnology.com.au

Phone: (03) 98306288

AC CURRENT MEASUREMENT

1) Insert the black test lead into the negative COM jack.

2) For current measurements up to 4000

µ

A AC, set the function switch to the

µ

A position and insert the red

test lead banana plug into the

µ

A jack.

3) For current measurements up to 400mA, set the function switch to the mA position and insert the red test lead banana plug into the mA jack.

4) For current measurements up to 10A AC, set the function switch to the 10A position and insert the red test lead banana plug into the 10A jack.

5) Use the MODE button to select AC or DC Current.

"AC" must appear on the display.

6) Remove power from the circuit under test, then open the point in the circuit where you wish to measure current.

7) Touch the black test probe tip to the negative side of the circuit. And touch the red test probe tip to the positive side of the circuit.

8) Apply power to the circuit.

9) Read the AC current measurement on the display.

www.procontechnology.com.au

Phone: (03) 98306288

RESISTANCE [

] MEASUREMENT

WARNING:

To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any resistance measurements.

1) Set the function switch to the

position.

2) Insert the black test lead into the negative COM jack and the red test lead plug into the positive

jack.

Press the MODE button until "M Ω " appears on the display.

3) Touch the test probe tips across the circuit or part under test. It is best to disconnect one side of the part under test so the rest of the circuit will not interfere with the reading.

4) Read the resistance measurement on the display.

CONTINUITY CHECK

WARNING: To avoid electric shock, disconnect

power to the circuit under test and discharge all capacitors before taking any continuity measurements.

www.procontechnology.com.au

Phone: (03) 98306288

1) Set the function switch to the position.

2) Insert the black lead plug into the COM jack and the red test lead plug into the positive jack.

3) Press the MODE button until " " appears on the display.

4) Touch the test probe tips to the circuit or wire you wish to check.

5) If the resistance is less than 50 Ω , an audible signal will sound. The display will also show the actual resistance measured between 0 Ω and 399.9

DIODE TEST

WARNING:

To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any diode measurements.

1) Set the function switch to the position.

2) Insert the black test lead into the COM jack and the red test lead into the jack.

3)

Press the MODE button until " " appears on the display.

4) Touch the test probe tips to the diode or

www.procontechnology.com.au

Phone: (03) 98306288

semiconductor junction you wish to test. It is best to disconnect one side of the part under test so the rest of the circuit will not interfere with the readings.

5) Reverse the probe polarity by switching probe positions. Note both readings.

6) The diode or junction can be evaluated as follows:

A. If one reading shows a value and the other reading shows OL, the diode is good.

B. If both readings show OL, the device is open.

C. If both readings are very small or zero, the device is shorted.

NOTE: The value indicated on the display during the

diode test is the forward voltage of the device.

CAPACITANCE MEASUREMENT

WARNING: To avoid electric shock, discharge the capacitor under test before making measurements.

1) Set the function switch to the CAP capacitance position. Press the MODE button until "nF" appears on the display.

2) Insert the black test lead into the negative COM jack and the red test lead into the positive CAP jack.

www.procontechnology.com.au

Phone: (03) 98306288

3) Touch the test probes across the part under test.

4) Read the capacitance value on the display.

FREQUENCY MEASUREMENT

1) Set the function switch to the Hz% position.

2) Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the positive Hz% jack.

3) Touch the test probe tips to the circuit under test.

4) Read the frequency on the display.

TEMPERATURE MEASUREMENT

1) Set the function switch to the Temp °C °F position.

2) Insert the Temperature Probe into the input jacks, making sure to observe the correct polarity.

3) Press the MODE button to select "°C" or "°F" on the display.

4) Touch the head of the thermocouple to the item whose temperature you wish to measure. Leave the thermocouple head touching the part until the reading stabilizes (about 30 seconds).

5) Read the temperature on the display.

www.procontechnology.com.au

Phone: (03) 98306288

Note: The temperature probe is fitted with a type K mini connector. A mini connector to banana connector adaptor is supplied for connection to the input jacks.

PC WIRELESS COMMUNICATION:

1) Install the USB wireless adapter, and launch the software supplied.

2) Press USB to enter the RF wireless transmit mode.

3) The RF icon " " will appear on the display.

4) When communication is established, the led indicator on the receiver will blink. If not, "initialize" the USB wireless receiver from the software menu selection.

5) Up to 3 times per second, the data will be displayed on the PC screen, plotted and inserted into a list.

6) Press USB to exit the RF wireless transmit mode.

www.procontechnology.com.au

Phone: (03) 98306288

SPECIFICATIONS

Technical Specifications:

Insulation: Class2, Double insulation.

Overvoltage category: CATIV 600V, CAT III 1000V

NOTE: This meter meets CAT III and CAT IV IEC 61010 standards. The IEC 61010 safety standard defines four overvoltage categories (CAT I to IV) based on the magnitude of danger from transient impulses. CAT III meters are designed to protect against transients in fixed-equipment installations at the distribution level; CAT IV meters are designed to protect against transients from the primary supply level (overhead or underground utility service).

Maximum voltage between any terminal and earth ground:

1000V DC/AC RMS

Display: 4000 counts LCD display

Polarity: Automatic, (-) negative polarity indication.

Over-range: "OL" indication.

AC Response: True RMS

Low battery indication: A battery " " symbol is displayed when the battery voltage drops below the operating level.

Auto power off: Meter automatically shuts down after approx. 30 minutes of inactivity unless in wireless mode.

www.procontechnology.com.au

Phone: (03) 98306288

Operating environment: -10°C to 50°C (14°F to 122°F) at <70% relative humidity.

Storage temperature: -30°C to 60°C (-4°F to 140°F) at <80% relative humidity

Relative humidity:

Maximum altitude:

90% (0°C to 30°C); 75% (30 o

C to

40°C); 45% (40 o

C to 50°C)

Operating: 3000m

Storage: 10,000m

Pollution degree: 2

Safety: The instrument complies with IEC/EN 61010-1:2001

Power:

and IEC/EN 61010-031:2002

9V battery, NEDA 1604, IEC 6F22

Dimensions:

182 (H) x 82 (W) x55 (D) mm

Weight:

Enclosure:

Approx. 380g.

Double moulded, waterproof (IP67)

Accuracy is given at 18°C to 28°C (65°F to 83°F

),

< 70% RH

DC Voltage (Auto-ranging)

Range Resolution Accuracy

400.0mV 0.1mV

4.000V 1mV

±0.8% of reading

40.00V 10mV

400.0V 100mV

±2 digits

1000V 1V

±1% of rdg ±2 digits

www.procontechnology.com.au

Phone: (03) 98306288

Input Impedance: 10M Ω

Maximum Input:

1000V DC

or

1000V AC

RMS

AC Voltage

(Auto-ranging)

Range Resolution Accuracy

400.0mV 0.1mV

4.000V 1mV

40.00V 10mV

400.0V 100mV

1000V 1V

±1.0% of reading

±3 digits

±1.2% of rdg ±5 digits

All AC voltage ranges are specified

from 5% of range to 100% of range

Input Impedance: 10M Ω

AC Response: 50Hz to 400Hz

Maximum Input: 1000V DC or 1000V AC RMS

DC Current (Auto-ranging)

Range Resolution Accuracy

400.0µA 0.1µA

4000µA 1µA

40.00mA 10µA

400.0mA 100µA

±1.2% of reading

±3 digits

10A 10mA ±2.5% of reading

±3 digits

www.procontechnology.com.au

Phone: (03) 98306288

Overload Protection: FF500mAI1000V and F10A/1000V

Maximum Input: 400µA DC on µA range

400mA DC on mA range

10A DC on 10A range

AC Current (Auto-ranging)

Range Resolution Accuracy

400.0µA 0.1µA

4000µA 1µA

40.00mA 10µA

400.0mA 100µA

10A 10mA

±1.5% of rdg ±5 digits

±3.0% of rdg ±5 digits

All AC Current ranges are specified from 5% of range to 100% of range

Overload Protection: FF500mA/1000V and F10A

I

1000V

AC Response: 50Hz to 400Hz

Maximum Input: 400µA AC RMS on µA

400mA AC RMS on mA

10A AC RMS on 10A range

www.procontechnology.com.au

Phone: (03) 98306288

Resistance [ Ω

] (Auto-ranging)

Range Resolution Accuracy

400.00

Ω 0.1

Ω ±0.8% of rdg ±5 digits

4.000k

Ω 1 Ω

40.00k

Ω 10 Ω

400.0k

Ω 100 Ω

4.000M

Ω 1k Ω

40.00M

Ω 10k Ω

±0.8% of rdg ±2 digits

±2.5% of rdg ±8 digits

Input Protection: 1000V DC or 1000V AC RMS

Capacitance (Auto-ranging)

Range Resolution Accuracy

40.00nF 10pF ±5.0% of rdg ±7 dgts

400.0nF 0.1nF

4.000µF 1nF ±3.0% of rdg ±5 dgts

40.00µF 10nF

100.0µF 0.1µF ±5.0% of rdg ±7 dgts

Input Protection: 1000V DC or 1000V AC RMS

Frequency (Auto-ranging)

Range Resolution Accuracy

4.000Hz 0.001Hz

www.procontechnology.com.au

Phone: (03) 98306288

40.00Hz 0.01Hz

400.0Hz 0.1Hz

4.000kHz 1Hz

40.00kHz 10Hz

400.0kHz 100Hz

5.00MHz 1kHz

±1.0% of rdg ±3 dgts

±1.2% of rdg ±4 dgts

Sensitivity: >0.5V RMS when

1MHz

Sensitivity: >3V RMS when >1MHz

Input Protection: 1000V DC or 1000V AC RMS

Duty Cycle

Range Resolution Accuracy

0.1%~99.9% 0.1% ±1.2% of rdg ±2 dgts

Pulse width:

Frequency:

>100us, <1 00ms

5Hz to 150kHz

Sensitivity: <0.5V RMS

Overload protection: 1000V DC or AC RMS

Temperature

Range Resolution Accuracy

-20 o

C ~ +760°C 1°C ±3% of rdg ±5 dgts

-4°F ~ +1400°F 1°F ±3% of rdg ±8 dgts

Sensor: Type K Thermocouple

www.procontechnology.com.au

Phone: (03) 98306288

Overload protection: 1000V DC or AC RMS

Diode Test

Test current

1mA typical

Max. 1.5V open

Resolution Accuracy

1mV ±10% of reading

±5 digits

Open circuit voltage: Max. 1.5V DC

Overload protection: 1000V DC or AC RMS

Audible continuity

Audible threshold: Less than 50 Ω

Test current max. 1.5mA

Overload protection: 1000V DC or AC RMS

Accessories

Included accessories

1) Standard Red/Black lead set with test probes.

2) K type thermocouple with adaptor.

3) USB wireless receiver with software for data logging.

4) Carry pouch.

www.procontechnology.com.au

Phone: (03) 98306288

BATTERY and FUSE replacement

WARNING:

To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery cover or opening the meter.

1) When the battery becomes exhausted or drops below the operating voltage, the battery warning " " will appear on the LCD display. The battery should be replaced.

2) Follow the instructions for installing a new battery.

See the Battery Installation instructions below.

3) Dispose of the old battery properly.

NOTE: Procon Technology has discovered that the

battery warning indication does not work. The best way to detect a low battery is to turn on the backlight and when this becomes very dim then the battery should be replaced. Note: a low battery is detected using the software available from Procon Technology. This is indicated on-screen and in the logged data as "Batt".

www.procontechnology.com.au

Phone: (03) 98306288

BATTERY INSTALLATION

WARNING:

To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery cover.

1) Disconnect the test leads from the meter.

2) Open the battery cover by removing the two central screws using a Phillips head screwdriver.

3) Remove the cover and insert the battery into battery compartment, observing the correct polarity.

4) Put the battery cover back in place. Secure with the two screws.

NOTE:

If your meter does not work properly, check the fuses and the battery to ensure that they are still good and that they are properly inserted.

REPLACING THE FUSES

WARNING:

To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery cover or opening the meter.

1) Disconnect the test leads from the meter and any item under test.

2. Remove the battery cover and battery by removing the

www.procontechnology.com.au

Phone: (03) 98306288

two central screws using a Phillips head screwdriver.

3) Remove the pull-out stand and remove the six screws using a good-quality Phillips head screwdriver. Be careful not to lose the screws, washers and rubber seals.

4) Remove the back of the meter with care.

5) Remove the old fuse from its holder by gently pulling it out. If necessary, use a screwdriver as a lever.

6) Install the new fuse into the holder.

7) Always use a fuse of the proper size and value.

(0.5A/1000V fast blow for the 400mA range [SIBA

70-172-40], 10A/1000V fast blow for the 10A range

[SIBA 50-199-06]).

8) Replace and secure the rear cover, battery and battery cover.

WARNING:

To avoid electric shock, do not operate the meter until the rear cover and battery cover are in place and fastened securely.

www.procontechnology.com.au

Phone: (03) 98306288

advertisement

Related manuals

advertisement