Fluke 77 IV Digital Multimeter manual

Add to my manuals
22 Pages

advertisement

Fluke 77 IV Digital Multimeter manual | Manualzz
Model 77 Series IV
Digital Multimeter
Users Manual
PN 2695884
September 2006 Rev.1, 11/15
© 2006-2015 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies.
Lifetime Limited Warranty
Each Fluke 20, 70, 80, 170 and 180 Series DMM will be free from defects in material and workmanship for its lifetime. As used herein, “lifetime” is defined as seven years after Fluke discontinues manufacturing the product, but the warranty period shall be at least ten years from
the date of purchase. This warranty does not cover fuses, disposable batteries, damage from neglect, misuse, contamination, alteration,
accident or abnormal conditions of operation or handling, including failures caused by use outside of the product’s specifications, or normal
wear and tear of mechanical components. This warranty covers the original purchaser only and is not transferable.
For ten years from the date of purchase, this warranty also covers the LCD. Thereafter, for the lifetime of the DMM, Fluke will replace the
LCD for a fee based on then current component acquisition costs.
To establish original ownership and prove date of purchase, please complete and return the registration card accompanying the product, or
register your product on http://www.fluke.com. Fluke will, at its option, repair at no charge, replace or refund the purchase price of a defective
product purchased through a Fluke authorized sales outlet and at the applicable international price. Fluke reserves the right to charge for
importation costs of repair/replacement parts if the product purchased in one country is sent for repair elsewhere.
If the product is defective, contact your nearest Fluke authorized service center to obtain return authorization information, then send the
product to that service center, with a description of the difficulty, postage and insurance prepaid (FOB Destination). Fluke assumes no risk
for damage in transit. Fluke will pay return transportation for product repaired or replaced in-warranty. Before making any non-warranty repair, Fluke will estimate cost and obtain authorization, then invoice you for repair and return transportation.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE
EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY. AUTHORIZED RESELLERS ARE NOT
AUTHORIZED TO EXTEND ANY DIFFERENT WARRANTY ON FLUKE’S BEHALF. Since some states do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you. If any provision of this
warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity
or enforceability of any other provision.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
Visit the Fluke website at: www.fluke.com
Register your Meter at: register.fluke.com
2/02
Table of Contents
Title
Page
How to Contact Fluke .............................................................................................................. 1
Safety Information .................................................................................................................... 1
Symbols ................................................................................................................................... 2
Unsafe Voltage ........................................................................................................................ 3
Test Lead Alert ........................................................................................................................ 3
Battery Saver (Sleep Mode) .................................................................................................... 4
Terminals ................................................................................................................................. 4
Rotary Switch Positions ........................................................................................................... 4
Display ..................................................................................................................................... 5
MIN MAX AVG Recording Mode ............................................................................................. 6
AutoHOLD Modes .................................................................................................................... 6
YELLOW Button ...................................................................................................................... 6
Display Backlight ..................................................................................................................... 6
Manual Ranging and Autoranging ........................................................................................... 7
Power-Up Options ................................................................................................................... 7
Making Basic Measurements .................................................................................................. 8
Measuring AC and DC Voltage .......................................................................................... 8
Measuring Resistance ........................................................................................................ 8
Measuring Capacitance ...................................................................................................... 8
Testing for Continuity .......................................................................................................... 9
Testing Diodes .................................................................................................................... 9
Measuring AC or DC Current ............................................................................................. 10
Measuring Frequency ......................................................................................................... 10
Using the Bar Graph ................................................................................................................ 11
Cleaning ................................................................................................................................... 11
Testing the Fuses .................................................................................................................... 11
Replacing the Battery and Fuses ............................................................................................ 12
General Specifications ............................................................................................................. 13
i
ii
Model 77 Series IV
Digital Multimeter
The Fluke Model 77 Series IV is a battery-powered, average
responding-rms indicating multimeter (hereafter "the Meter"), with
a 6000-count, 3 3/4-digit display, and a bar graph.
How to Contact Fluke
To contact Fluke, call one of the following telephone numbers:
•
Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
•
Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
•
Canada: 1-800-36-FLUKE (1-800-363-5853)
•
Europe: +31 402-675-200
•
Japan: +81-3-6714-3114
•
Singapore: +65-6799-5566
• Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com.
To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals.
Safety Information
A Warning identifies conditions and procedures that are
dangerous to the user. A Caution identifies conditions and
procedures that can cause damage to the Product or the
equipment under test.
XW Warning
To prevent possible electrical shock, fire, or personal
injury.
• Use the Product only as specified, or the protection
supplied by the Product can be compromised.
• Comply with local and national safety codes. Use
personal protective equipment (approved rubber
gloves, face protection, and flame-resistant clothes)
to prevent shock and arc blast injury where
hazardous live conductors are exposed.
• Use Product-approved measurement category (CAT),
voltage, and amperage rated accessories (probes,
test leads, and adapters) for all measurements.
• Remove all probes, test leads, and accessories that
are not necessary for the measurement.
• Do not use test leads if they are damaged. Examine
the test leads for damaged insulation and measure a
known voltage.
1
Model 77 Series IV
Users Manual
• Do not use a current measurement as an indication
that a circuit is safe to touch. A voltage measurement
is necessary to know if a circuit is hazardous.
• Do not use the Product if it is damaged.
• Disable the Product if it is damaged.
• Do not use the Product above its rated frequency.
• Do not use in CAT III or CAT IV environments without
the protective cap installed on test probe. The
protective cap decreases the exposed probe metal to
<4 mm. This decreases the possibility of arc flash
from short circuits.
• Replace a blown fuse with exact replacement only for
continued protection against arc flash.
• Do not use the HOLD function to measure unknown
potentials. When HOLD is turned on, the display
does not change when a different potential is
measured.
• Attempting to make a measurement with a test lead in
an incorrect terminal might blow a fuse, damage the
Meter, and cause serious personal injury.
Symbols
Description
Symbol
AC (Alternating Current)

Fuse

DC (Direct Current)
Conforms to European Union directives


DC/AC
P
)

W
2
Description
Symbol

Certified by CSA Group to North American
safety standards.
Earth
T
Double insulated
WARNING. RISK OF DANGER.
X
WARNING. HAZARDOUS VOLTAGE. Risk of
electric shock.
Digital Multimeter
Unsafe Voltage
Symbol
Description
Symbol
Description


Battery (Low battery when shown on display)

Consult user documentation.
Certified by TÜV SÜD Product Service.
Ã
Conforms to relevant South Korean EMC
Standards

Conforms to relevant Australian EMC standards
CAT II
Measurement Category II is applicable to test and measuring circuits connected directly to utilization points (socket
outlets and similar points) of the low-voltage MAINS installation.
CAT III
Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the
building’s low-voltage MAINS installation.
CAT IV
Measurement Category IV is applicable to test and measuring circuits connected at the source of the building’s lowvoltage MAINS installation.
Unsafe Voltage
Test Lead Alert
During a voltage measurement, the Meter alerts you to the
presence of a potentially hazardous voltage. When the Meter
detects a voltage ≥30 V or a voltage overload (OL), the Y symbol
is displayed.
To remind you to check that the test leads are in the correct
terminals,  is momentarily displayed when you move the
rotary switch to or from the mA or A position.
3
Model 77 Series IV
Users Manual
Battery Saver (Sleep Mode)
Rotary Switch Positions
The Meter enters the "Sleep" mode and blanks out the display if
there is no function change or button press for 20 minutes. To
disable the Sleep mode, hold down the yellow button while turning
the Meter on. The Sleep mode is always disabled in the MIN MAX
AVG mode and the AutoHOLD mode.
Terminals
Switch
Position

Hz

.
1
3
m
e
V

FUSED
2
4
R


AIK01F.EPS
Item
1
2
3
4
4
Description
Input terminal for AC and DC milliamp measurements to
400 mA.
Input terminal for AC and DC current measurements to
10 A.
Input terminal for voltage, continuity, resistance, diode
test, capacitance, and frequency measurements.
Common (return) terminal for all measurements.
Measurement Function
AC voltage from 0.001 to 1000 V.
Frequency from 2 Hz to 99.99 kHz.
DC voltage from 1 mV to 1000 V.
DC mV from 0.1 mV to 600 mV.
Ohms from 0.1 Ω to 50 MΩ.
Farads from 1 nF to 9999 μF.
Beeper turns on at <25 Ω and turns off at >250 Ω.
Diode test. Displays OL above 2.4 V.
AC mA from 0.01 mA to 400 mA.
DC mA from 0.01 mA to 400 mA.
mA

A
AC A from 0.001 A to 10 A.
DC A from 0.001 A to 10 A
>10.00 display flashes.
>20 A, OL is displayed.
Digital Multimeter
Display
Display
No.
.
5
6
4
7
3
8
2
12
14
10
11
15
AIK02F.EPS
No.
Symbol

1
2
3

4
Y
5



6
7
8
DC, AC
Direct current, alternating current.
9

Low battery. Replace battery.
10
11
610000mV
Bar graph
Auto Range
All possible ranges.
Analog display.
The Meter selects the range with
the best resolution.
The user selects the range.
Bar graph polarity.
Manual Range
9
13
MAX , MIN,
AVG
nμ F, mVA,
Mke, kHz
Meaning
12
7
1
Symbol
13
±
14

15

Meaning
Continuity test.
Diode test.
Negative readings.
Unsafe voltage. Voltage ≥30 V, or
voltage overload (OL)
AutoHOLD is enabled. Display holds
present reading until it detects new
stable input. Then the Meter beeps
and displays new reading.
MIN MAX AVG enabled.
Maximum, minimum, average, or
present reading.
The input out of range.
WTest lead alert. Displayed when
the rotary switch is moved to or
from the mA or A position.
Error Messages


Replace the battery immediately.
In the capacitance function, too much electrical
charge is present on the capacitor being tested.


Invalid EEPROM data. Have Meter serviced.


Invalid calibration data. Calibrate Meter.
Measurement units.
5
Model 77 Series IV
Users Manual
MIN MAX AVG Recording Mode
AutoHOLD Modes
The MIN MAX AVG recording mode captures the minimum and
maximum input values, and calculates a running average of all
readings. When a new high or low is detected, the Meter beeps.
XWWarning
To avoid electric shock, do not use the AutoHOLD mode
to determine if a circuit is live. Unstable or noisy
readings will not be captured.
In the AutoHOLD mode, the Meter holds the reading on the display
until it detects a new stable reading. Then the Meter beeps, and
displays the new reading.
• Press HOLD to activate AutoHOLD.  lights.
• Press HOLD again or turn the rotary switch to resume normal
operation.
Note
For dc functions, accuracy is the specified accuracy of the
measurement function ±12 counts for changes longer than
350 ms in duration.
For ac functions, accuracy is the specified accuracy of the
measurement function ±40 counts for changes longer than
350 ms in duration.
To use MIN MAX AVG recording:
• Make sure that the Meter is in the desired measurement
function and range. (Autoranging is disabled in the MIN MAX
AVG mode.)
• Press MIN MAX to activate MIN MAX AVG mode.
 and MAX light, and the highest reading detected since
entering MIN MAX AVG is displayed.
• Press MIN MAX to step through the low (MIN), average (AVG),
and present readings.
• To pause MIN MAX AVG recording without erasing stored
values, press HOLD.  is displayed.
• To resume MIN MAX AVG recording, press HOLD again.
 turns off.
• To exit and erase stored readings, press MIN MAX for 1 second
or turn the rotary switch.
6
YELLOW Button
Press the yellow button to select alternate measurement functions
on a rotary switch setting, e.g., to select DC mA, DC A, Hz,
capacitance, or diode test.
Display Backlight
Press  to toggle the backlight on and off. The backlight
automatically turns off after 2 minutes.
Digital Multimeter
Manual Ranging and Autoranging
Manual Ranging and Autoranging
Power-Up Options
The Meter has both Manual range and Autorange modes.
• In the Autorange mode, the Meter selects the range with the
best resolution.
• In the Manual Range mode, you override Autorange and select
the range yourself.
When you turn the Meter on, it defaults to Autorange and Auto
Range is displayed.
1. To enter the Manual Range mode, press RANGE.
Manual Range is displayed.
2. In the Manual Range mode, press RANGE to increment the
range. After the highest range, the Meter wraps to the lowest
range.
To select a Power-Up Option, hold down the button indicated while
turning the Meter on.
Power-Up Options are cancelled when the Meter is turned OFF.
Note
You cannot manually change the range in the MIN MAX
AVG mode.
If you press RANGE while in MIN MAX AVG, the Meter
beeps twice, indicating an invalid operation, and the range
does not change.
3. To exit Manual Range, press RANGE for 1 second or turn the
rotary switch.
The Meter returns to Autorange and Auto Range is displayed.
Button
Power-Up Options

Turns on all display segments when in VAC switch
position.

Disables beeper.  is diplayed when enabled.


(YELLOW)

Enables "Smoothing" mode.  is displayed when
enabled.
Dampens display fluctuations of rapidly changing
inputs by digital filtering.
Disables automatic power-down ("Sleep mode").
 is displayed when enabled
Sleep mode is also disabled while the Meter is in a
MIN MAX AVG Recording mode, or the AutoHOLD
mode.
Disables automatic 2-minute backlight timeout.
 is displayed when enabled.
7
Model 77 Series IV
Users Manual
Measuring Resistance
Making Basic Measurements
The figures on the following pages show how to make basic
measurements.
When connecting the test leads to the circuit or device, connect
the common (COM) test lead before connecting the live lead;
when removing the test leads, remove the live lead before
removing the common test lead.
HOLD
MIN MAX
RANGE
XWWarning
To avoid electric shock or injury, or damage to the
Meter, disconnect circuit power and discharge all highvoltage capacitors before testing resistance, continuity,
diodes, or capacitance.
Measuring AC and DC Voltage
Volts AC
Volts DC
Millivolts DC
AIK04F.EPS
Measuring Capacitance
HOLD
V
MIN MAX
RANGE
HOLD
MIN MAX
RANGE
HOLD
MIN MAX
RANGE
mV
V
HOLD
+
_
+
MIN MAX
RANGE
_
AIK03F.EPS
_
+
+
AIK05F.EPS
8
Digital Multimeter
Making Basic Measurements
Testing for Continuity
Testing Diodes
Good Diode
HOLD
MIN MAX
RANGE
HOLD
MIN MAX
Good Diode
RANGE
HOLD
MIN MAX
RANGE
HOLD
_
+
MIN MAX
RANGE
_
+
Single Beep
AIK06F.EPS
Forward Bias
Reverse Bias
Bad Diode
Bad Diode
HOLD
MIN MAX
RANGE
HOLD
MIN MAX
RANGE
CAT
_
+
_
+
and
Open
Shorted
AIK07F.EPS
9
Model 77 Series IV
Users Manual
Measuring AC or DC Current
Measuring Frequency
XWWarning
To avoid personal injury or damage to the Meter:
• Never attempt to make an in-circuit current measurement when the open-circuit potential to earth ground
is >1000 V.
• Check the Meter's fuses before testing. (See “Testing
the Fuses”.)
• Use the proper terminals, switch position, and range
for your measurement.
• Never place the probes in parallel with a circuit or
component when the leads are plugged into the
current terminals.
Turn power OFF, break circuit, insert Meter in series, turn power
on.
XWWarning
To avoid electrical shock, disregard the bar graph for
frequencies >1 kHz. If the frequency of the measured
signal is >1 kHz, the bar graph is unspecified.
The Meter measures the frequency of a signal. The trigger level is
0 V ac for all ranges.
AC Voltage Frequency
HOLD
MIN MAX
RANGE
Hz
V
DC
HOLD
MIN MAX
+
RANGE
mA
EOM09F.EPS
• To exit frequency, press yellow button or turn the rotary switch.
• In frequency, the bar graph shows the ac voltage accurately up
to 1 kHz.
• Select progressively lower ranges using manual ranging for a
stable reading.
A
+
CAT
CAT
+
AIK08F.EPS
10
Digital Multimeter
Using the Bar Graph
Using the Bar Graph
Cleaning
The bar graph is like the needle on an analog Meter. There is an
overload indicator (►) to the right, and a polarity indicator (±) to
the left.
Because the bar graph is much faster than the digital display, the
bar graph is useful for making peak and null adjustments, and for
observing rapidly changing inputs.
The bar graph is disabled when measuring capacitance. In
frequency, the bar graph accurately indicates the voltage or
current up to 1 kHz.
The number of lit segments indicates the measured value and is
relative to the full-scale value of the selected range.
For example, in the 60 V range (see below), the major divisions on
the scale represent 0, 15, 30, 45, and 60 V. An input of −30 V
lights the negative sign and the segments up to the middle of the
scale.
Wipe the case with a damp cloth and mild detergent. Do not use
abrasives or solvents. Dirt or moisture in the terminals can affect
readings.
Testing the Fuses
XWWarning
To avoid electrical shock or injury, remove the test
leads and any input signals before replacing the fuse.
Test fuses as shown below.
440 mA
<12 Ω
HOLD
MIN MAX
RANGE
11 A
OK <0.5
OK
OK
OK
V
AIK11F.EPS
AIK12F.EPS
11
Model 77 Series IV
Users Manual
Replacing the Battery and Fuses
XWWarning
To avoid shock, injury, or damage to the Meter:
• Use ONLY fuses with the amperage, interrupt,
voltage, and speed ratings specified. Disconnect test
leads before opening case.


B1
F1
F2
AIK13F.EPS
12
F1 Fuse, 440 mA, 1000 V, FAST
Fluke PN 943121
F2 Fuse, 11 A, 1000 V, FAST
Fluke PN 803293
B1 Battery, 9 V Alkaline,
NEDA 1604 / 1604A
Fluke PN 614487
Digital Multimeter
General Specifications
General Specifications
Accuracy is specified for 1 yr after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Accuracy
specifications take the form of ±([ % of Reading ] + [ Counts ]).
Maximum voltage between any
terminal and earth ground ............................................ 1000 V
Fuse protection for mA inputs ..................................... 0.44 A, 1000 V, IR 10 kA
Fuse protection for A inputs ........................................ 11 A, 1000 V, IR 17 kA FAST Fuse
Display
Digital .............................................................................. 6000 counts, updates 4/sec
Bar Graph ........................................................................ 33 segments; Updates 32/sec
Frequency ....................................................................... 10,000 counts
Capacitance .................................................................... 1,000 counts
Altitude
Operating .................................................................... 2,000 meters
Storage ........................................................................ 12,000 meters
Temperature
Operating .................................................................... -10 °C to +50 °C
Storage........................................................................ -40 °C to +60 °C
Temperature coefficient ................................................ 0.1 X (specified accuracy) / °C (< 18 °C or > 28 °C)
Safety
General ....................................................................... IEC 61010-1: Pollution Degree 2
Measurement .............................................................. IEC 61010-2-033: CAT IV 600 V / CAT III 1000 V
Relative Humidity (Maximum Non-condensing) ......... 90 % to 35 °C
75 % to 40 °C
45 % to 50 °C
13
Model 77 Series IV
Users Manual
Battery Life .................................................................... 400 hrs typical (Alkaline)
Size (H x W x L) ............................................................. 4.3 cm x 9 cm x 18.5 cm
Weight ............................................................................ 420 g
Electromagnetic Compatibility (EMC)
International ................................................................ IEC 61326-1: Portable Electromagnetic Environment, IEC 61326-2-2.
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductivelycoupled radio frequency energy that is necessary for the internal function of the
equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic
and those directly connected to a low-voltage power supply network that
supplies buildings used for domestic purposes. There may be potential
difficulties in ensuring electromagnetic compatibility in other environments due to
conducted and radiated disturbances.
Emissions that exceed the levels required by CISPR 11 can occur when the
equipment is connected to a test object.
Korea (KCC) ............................................................... Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave
equipment and the seller or user should take notice of it. This equipment is
intended for use in business environments and not to be used in homes.
USA (FCC) .................................................................. 47 CFR 15 subpart B. This product is considered an exempt device per clause
15.103.
14
Digital Multimeter
General Specifications
Function
Range
Resolution
Accuracy ± ( [ % of Reading ] + [ Counts ] )
6.000 V
60.00 V
600.0 V
1000 V
0.001 V
0.01 V
0.1 V
1V
2.0 % + 2
(45 Hz to 1 kHz)
DC mV
600.0 mV
0.1 mV
0.3 % + 1
DC Volts
6.000 V
60.00 V
600.0 V
1000 V
0.001 V
0.01 V
0.1 V
1V
0.3 % + 1
Continuity
600 Ω
1Ω
Ohms
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
50.00 MΩ
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
Meter beeps at <25 Ω, beeper turns off at >250 Ω;
detects opens or shorts of 250 μs or longer.
0.5 % + 2
0.5 % + 1
0.5 % + 1
0.5 % + 1
0.5 % + 1
2.0 % + 1
Diode test
2.400 V
0.001 V
1%+2
Capacitance
1000 nF
10.00 μF
100.0 μF
9999 μF[1]
1 nF
0.01 μF
0.1 μF
1 μF
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
AC Amps (Average
responding)[2]
60.00 mA
400.0 mA[3]
6.000 A
10.00 A[4]
0.01 mA
0.1 mA
0.001 A
0.01 A
2.5 % + 2
(45 Hz to 1 kHz)
AC Volts (Average
responding)
[1]
[2]
[3]
[4]
In the 9999 μF range for measurements to 1000 μF, the measurement accuracy is 1.2 % + 2.
Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.
400.0 mA accuracy specified up to 600 mA overload.
>10 A unspecified.
15
Model 77 Series IV
Users Manual
Function
DC Amps
[3]
Range
Resolution
Accuracy ± ( [ % of Reading ] + [ Counts ] )
60.00 mA
400.0 mA[4]
6.000 A
10.00 A[5]
0.01 mA
0.1 mA
0.001 A
0.01 A
1.5 % + 2
Hz [1][ 2]
(ac voltage input)
99.99 Hz
999.9 Hz
9.999 kHz
99.99 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 % + 1
MIN MAX AVG
For dc functions, accuracy is the specified accuracy of the measurement function ±12 counts for changes longer than
350 ms in duration.
For ac functions, accuracy is the specified accuracy of the measurement function ±40 counts for changes longer than
350 ms in duration.
[1]
[2]
[3]
[4]
[5]
16
Frequency is specified from 2 Hz to 99.99 kHz.
Below 2 Hz, the display shows zero Hz.
Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.
400.0 mA accuracy specified up to 600 mA overload.
>10 A unspecified.
Digital Multimeter
General Specifications
[1]
Function
Overload Protection
Input Impedance
(Nominal)
Common Mode Rejection Ratio
(1 kΩ Unbalanced)
Normal Mode Rejection
Volts AC
1000 V
>10 MΩ <100 pF
>60 dB @ dc, 50 Hz or 60 Hz
Volts DC
1000 V
>10 MΩ <100 pF
>120 dB @ dc, 50 Hz or 60 Hz
>60 dB @ 50 Hz or 60 Hz
mV
1000 V
>10 MΩ <100 pF
>120 dB @ dc, 50 Hz or 60 Hz
>60 dB @ 50 Hz or 60 Hz
Open Circuit Test
Voltage
Full Scale Voltage To:
6.0 MΩ
50 MΩ
Short Circuit Current
[2]
[2]
<8.0 V dc
[2]
<8.0 V dc
Ohms/Capacitance
1000 V
Continuity/Diode test
1000 V
[1]
[2]
<660 mV dc
<4.6 V dc
<1.1 mA
2.4 V dc
<1.1 mA
10 7 V-Hz maximum.
For circuits <0.3 A short circuit. 660 V for high energy circuits.
Function
Overload Protection
Overload
mA
Fused, 440 mA, 1000 V FAST Fuse
600 mA overload for 2 minutes maximum, 10
minutes rest.
A
Fused, 11 A, 1000 V FAST Fuse
20 A overload for 30 seconds maximum, 10
minutes rest.
17
Model 77 Series IV
Users Manual