Digital StereoLoc v4.3 DICOM Conformance Statement


Add to my manuals
33 Pages

advertisement

Digital StereoLoc v4.3 DICOM Conformance Statement | Manualzz

Addendum

European Commission Authorized Representative Change

1:

Change of the Authorized Representative

This document updates your product manuals to change the address of the European Commission

Authorized Representative (EC Rep). This change is effective on the first day of November 2015. The new address is:

Hologic Ltd.

Heron House Oaks Business Park

Crewe Road

Wythenshawe, Manchester

M23 9HZ, UK

Tel: +44 (0)161 946 2206

Fax: +44 (0)161 602 0995

Place this document with your product manuals for future reference.

MAN-04594 Revision 001, October 2015 Hologic

36 Apple Ridge Road, Danbury, CT 06810 (203) 207-4500

© 2015 Hologic, Inc. The information contained in this document is proprietary. Distribution is strictly prohibited.

Page 1 of 1

Digital StereoLoc II DSM 4.3

DICOM Conformance Statement

Part Number MAN-02018

Revision 004

September 2014

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

Table of Contents

1.0

OVERVIEW ........................................................................................................................................... 4

2.0

INTRODUCTION .................................................................................................................................. 5

2.1.

A

UDIENCE

............................................................................................................................................ 5

2.2.

R

EMARKS

............................................................................................................................................. 5

2.3.

D

EFINITIONS

, T

ERMS AND

A

BBREVIATIONS

......................................................................................... 5

2.4.

R

EFERENCES

........................................................................................................................................ 6

3.0

NETWORKING ..................................................................................................................................... 7

3.1.

I

MPLEMENTATION

M

ODEL

................................................................................................................... 7

3.1.1 Application Data Flow ......................................................................................................................... 7

3.1.2 Functional Definition of AE’s .............................................................................................................. 8

3.1.3 Sequencing of Real World Activities .................................................................................................. 11

3.2.

AE S

PECIFICATIONS

........................................................................................................................... 12

3.2.1 DSM Client AEs ................................................................................................................................ 12

3.2.2 Storage Server AE .............................................................................................................................. 22

3.3.

N

ETWORK

I

NTERFACES

...................................................................................................................... 23

3.3.1 Physical Network Interface ................................................................................................................ 23

3.3.2 Additional Protocols .......................................................................................................................... 23

3.4.

C

ONFIGURATION

................................................................................................................................ 23

3.4.1 AE Title/Presentation Address Mapping ............................................................................................ 24

3.4.2 Configurable Parameters for Local AEs ............................................................................................ 25

4.0

MEDIA INTERCHANGE ................................................................................................................... 25

5.0

SUPPORT OF CHARACTER SETS ................................................................................................. 25

6.0

SECURITY ........................................................................................................................................... 25

7.0

ANNEXES ............................................................................................................................................. 26

7.1.

IOD C

ONTENTS

.................................................................................................................................. 26

7.1.1

Created SOP Instance(s) .............................................................................................................. 26

7.1.2

Usage of Attributes from Received IODs ...................................................................................... 32

7.1.3

Attribute Mapping ......................................................................................................................... 32

7.1.4

Coerced/Modified Attributes ........................................................................................................ 32

7.2.

D

ATA

D

ICTIONARY OF

P

RIVATE

A

TTRIBUTES

.................................................................................... 32

7.3.

C

ODED

T

ERMINOLOGY AND

T

EMPLATES

........................................................................................... 32

7.4.

G

RAYSCALE

I

MAGE

C

ONSISTENCY

.................................................................................................... 32

7.5.

S

TANDARD

E

XTENDED

/S

PECIALIZED

/P

RIVATE

SOP C

LASSES

........................................................... 32

7.6.

P

RIVATE

T

RANSFER

S

YNTAXES

.......................................................................................................... 32

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 3 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

1.0 Overview

The Hologic Selenia Digital StereoLoc II Digital Spot Mammography (DSM) option implements the necessary DICOM services to send acquired Digital Mammography For Presentation stereotactic images to a networked storage device, request Storage Commitment from a networked storage device, query and retrieve Hologic Digital Mammography For Presentation stereotactic images from a networked storage device, and print to a networked hardcopy device.

The DSM is not a standalone product. It is an add-on to the Selenia Acquisition Workstation 3.3 or later.

Note: Refer to the Selenia Acquisition Workstation DICOM conformance statement for information regarding DICOM Modality Worklist SOP Class conformance for the DSM.

Table 1-1 provides an overview of the network services supported by the DSM when it is enabled on a Selenia Acquisition Workstation.

Table 1-1

Network Services

User of Service (SCU) Provider of Service (SCP) SOP Classes

Transfer

Digital Mammography X-Ray Image

Storage – For Presentation

Print Management

Basic Grayscale Print Management Meta

SOP Class

Workflow Management

Storage Commitment Push Model SOP

Class

Query/Retrieve

Study Root Query/Retrieve Information

Model - FIND

Study Root Query/Retrieve Information

Model – MOVE

Yes

Yes

Yes

Yes

Yes

Yes

No

No

No

No

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 4 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

2.0 Introduction

Hologic, Inc. develops and markets a full line of mammography products. The DSM provides the capability to utilize the amorphous selenium X-ray detector of the Selenia full-field digital mammography system for breast stereotactic imaging. The DSM provides a graphical user interface

(GUI) with a keyboard, pointing device (trackball), monitor, and DICOM interfaces separate from the

Selenia Acquisition Workstation.

2.1. Audience

This document contains the DICOM conformance claim for the DSM. This document is intended to aid in connecting the DSM to other components that make use of the DICOM standard for interconnecting networked imaging devices.

The information within this document applies to DSM. The reader of this document should be familiar with the DICOM standard and PACS components that utilize the standard.

2.2. Remarks

A DICOM conformance statement—the structure and content of which are stipulated by the DICOM standard—is intended to aid in determining the suitability of interconnecting digital imaging devices.

References to specific functionality in a conformance statement are not sufficient to guarantee interoperability between components. The following should be considered when evaluating interoperability:

• This conformance statement provides a starting point for ascertaining whether the product can communicate with other systems.

• The only way to know for certain whether the DSM can interoperate with other systems is to perform connectivity testing.

• This document represents a best effort to document the functionality of commercial versions of the DSM and is not a functional specification of any Hologic component or product. Hologic reserves the right to make changes at any time to the functionality of the DICOM components described herein, and is committed to following the evolution of the DICOM standard.

2.3. Definitions, Terms, and Abbreviations

Amorphous Selenium: Semiconductor material used in the direct capture X-ray detector of the

Selenia.

ACR: American College of Radiology

AE: Application Entity

DICOM: Digital Imaging and Communications in Medicine

DIMSE: DICOM Message Service Element

DSM: Selenia Digital StereoLoc II Digital Spot Mammography

GSDF: Grayscale Standard Display Function

IOD: Information Object Definition

LUT: Lookup Table

MWL: Modality Worklist

NEMA: National Electrical Manufacturers Association

PDU: Protocol Data Unit

SCP: Service Class Provider

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 5 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

SCU: Service Class User

SOP: Service Object Pair

TCP/IP: Transmission Control Protocol/Internet Protocol

VR: Value Representation

2.4. References

• NEMA PS3 / ISO 12052, Digital Imaging and Communications in Medicine (DICOM)

Standard, National Electrical Manufacturers Association, Rosslyn, VA, USA (available free at http://medical.nema.org/)

• Selenia Acquisition Workstation DICOM Conformance Statement – MAN-01287

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 6 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

3.0 Networking

3.1. Implementation Model

3.1.1 Application Data Flow

Select Patient on Selenia AWS

Print Screen or

Print Film

Request

File Manager

Copy Request

Auto Archive

Request

Query/Retrieve

Request

DICOM Echo

Modality Worklist

Client

Application Entity

Print Client

Application Entity

Storage Client

Application Entity

Storage

Commitment

Client

Application Entity

Query/Retrieve

Client

Application Entity

Storage Server

Application Entity

Verification

Client

Application Entity

Remote

Application Entity

Provides Worklist

Items

Remote

Application Entity

Prints Film Sheets

Remote

Application Entity

Stores Images

Remote

Application Entity

Commits Images

Remote

Application Entity

Responds to

Query/Retrieve

Request

Remote

Application Entity

Sends Images

Remote

Application Entity

For Verification

DICOM Standard Interface

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 7 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

3.1.1.1 Modality Worklist Client

Refer to the Selenia Acquisition Workstation DICOM Conformance Statement for information regarding the DICOM Modality Worklist SOP Class.

3.1.1.2 Print Client

• The DSM provides two methods for printing hardcopy:

Print Screen: The user selects the Contents (Left image, Right image, Image pair, Entire screen), Destination (DICOM Printer), and Number of copies (1 – 99, default = 1). The selected portion of the display screen is captured, and each film is printed as one-up (1x1). Left image and

Right image are printed portrait orientation, while Image pair and Entire screen are printed landscape orientation.

Print Film: The user has selected one or more images from a list prior to selecting Print Film.

From a dialog, the user selects the Destination (DICOM Printer), Film format (1x1, 2x2, 3x2, 4x3),

Print resolution (Hi res), and Number of copies (1 – 99, default = 1). The images are prepared for printing, and all films are printed portrait orientation.

3.1.1.3 Storage and Storage Commitment Client

The DSM provides two methods for sending images to a remote DICOM storage server:

File Manager Copy: The user selects one or more images from the list, selects a remote DICOM storage server as the destination, and selects Copy. The images are transmitted via network to the remote DICOM storage server. Up to three remote DICOM storage servers may be configured during Setup.

Auto Archive: During the DSM shut down procedure (if configured), or by user request from the main panel, all images that have not been archived are copied to the configured Auto Archive destination. A remote DICOM storage server may be configured as the Auto Archive destination during Setup. If the remote DICOM storage server supports Storage Commitment, DSM images are marked as archived after successful Storage and Storage Commitment. If not, DSM images are marked as archived after successful Storage, but there is no guarantee that the remote

DICOM storage server will keep the images long-term.

3.1.1.4 Query/ Retrieve Client

In the DSM DICOM Query dialog, the user may enter a Patient ID or Patient Name, and optionally a

Study Date, then selects a remote DICOM query server. Study and series level queries are performed, with Modality set to “MG”. The query results are presented to the user in a DICOM

Retrieve dialog. The user may select one or more series for which to retrieve the associated images.

3.1.1.5 Verification Client

Selecting the DICOM Echo button in the DICOM Print Setup and DICOM Server Setup dialogs, accessed from the Setup menu, sends a Verification request to the associated remote DICOM Print,

Storage, or Query/Retrieve server. A message box indicates success or failure.

3.1.2 Functional Definition of AE’s

The Client AEs are finite modules within the DSM. They exist only one at a time, for the duration of a single Association.

3.1.2.1 Functional Definition of the Modality Worklist Client Application Entity

Refer to the Selenia Acquisition Workstation DICOM Conformance Statement for information regarding the Modality Worklist Client Application Entity.

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 8 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

3.1.2.2 Functional Definition of Print Client Application Entity

After receiving input from the DSM user regarding images to print and the associated configuration, the Print Client performs the following DICOM communications, making use of the Basic Grayscale

Print Management Meta SOP Class by means of DIMSE services. The user is informed of progress via dialog:

• Attempt to establish an Association with the selected DICOM Printer, requesting Basic Grayscale

Print Management Meta.

• Request the current Printer status. Report to the user if the status is warning or failure and give the user the opportunity continue the print request, or release the Association.

• Depending on the number of images and format selected, create a Film Session with one or more

Film Boxes, each Film Box containing one or more Basic Grayscale Image Boxes. Each Basic

Grayscale Image Box contains the preformatted image pixel data for a single position within a

Film Box.

• Request action (print) for each Film Box.

• Delete the Film Session, and release the Association.

3.1.2.3 Functional Definition of Storage and Storage Commitment Client Application Entity

After receiving input from the user regarding images to copy, the Storage Client performs the following DICOM communications, making use of the Digital Mammography X-Ray Image Storage –

For Presentation SOP Class by means of DIMSE services. The user is informed of errors via dialog:

• Attempt to establish an Association with the selected remote DICOM storage server, requesting

Digital Mammography X-Ray Image Storage – For Presentation.

• Add DSM Private Attributes to the DICOM Data Dictionary.

• For each image to be stored, fill a composite DICOM Digital Mammography X-Ray Image IOD structure and send a C-STORE request.

• Release the Association after receiving the C-STORE response.

For Auto Archive, after successful DICOM storage, the following additional DICOM communications may be performed by the Storage Commitment Client, making use of the Storage Commitment Push

Model SOP Class by means of DIMSE services. The user is informed of progress via dialog:

• Attempt to establish an Association with the configured remote DICOM storage server, requesting Storage Commitment Push Model.

• Send an N-ACTION request, providing a list of successfully stored image SOP instance UIDs in the request.

• Release the Association after receiving the N-ACTION response.

• If the N-ACTION request was successful, wait for an Association request from the remote DICOM storage server, expecting an N-EVENT-REPORT request after Association acceptance. This message contains the results of the storage commitment request.

• Wait for an Association release.

• The Storage Commitment Client times out and stops waiting after the configured commit timeout if the Association request, N-EVENT-REPORT request, or Association release is not received.

The transaction UID is no longer valid after this occurs.

• Each successfully committed image is marked as archived on the DSM.

Note: If Storage Commitment is not configured, the image is marked as archived immediately after the Storage association is released.

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 9 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

3.1.2.4 Functional Definition of Query/ Retrieve Client Application Entity

Upon receiving input from the user regarding query and then retrieve parameters, the Query/Retrieve

Client performs the following DICOM communications, making use of the Study Root Query/Retrieve

Information Model – FIND and Study Root Query/Retrieve Information Model – MOVE SOP Classes, respectively, by means of DIMSE services. The user is informed of errors via dialog:

• Attempt to establish an Association with the selected remote DICOM query server, requesting

Study Root Query/Retrieve Information Model – FIND and Study Root Query/Retrieve

Information Model – MOVE.

• Send a Study level C-FIND request, with the Patient ID or Patient Name, and optionally Study

Date as input.

• If there are no matches or after the last match is received, release the Association.

• For each study returned, send a Series level C-FIND request, with Modality set to “MG”.

• Report the series level results to the user in a dialog. If there are no matches or after the last match is received, release the Association.

• For each selected series, send a Series level C-MOVE request, with a single Series UID as input.

• Release the Association after receiving the final C-MOVE response.

The Storage Server performs the following DICOM communications, making use of the Digital

Mammography X-Ray Image Storage – For Presentation SOP Class by means of DIMSE services.

Errors are recorded in a text output file:

• Wait for an Association request from the remote DICOM query server, accepting Digital

Mammography X-Ray Image Storage – For Presentation.

• Wait for and process C-STORE requests. Received images are stored on the DSM.

• Wait for an Association release.

• The Storage Server continues to wait for new Association requests.

3.1.2.5 Functional Definition of Verification Client Application Entity

Upon user request of DICOM Echo, the Verification Client performs the following DICOM communications, making use of the Verification SOP Class by means of DIMSE services. The user is informed of success or errors via dialog:

• Attempt to establish an Association with the remote DICOM Print, Storage, or Query/Retrieve server, requesting Verification.

• Send a C-ECHO request.

• Release the Association after receiving the C-ECHO response.

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 10 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

3.1.3 Sequencing of Real World Activities

Print: The order of DIMSE service requests is significant. A Film Session is created before the associated Film Box(es), and each Film Box is created before the associated Basic Grayscale Image

Box(es).

Storage Commitment: The N-EVENT-REPORT is expected in a separate association from the N-

ACTION request. The DSM releases the original Association and starts listening for an Association request after receiving a successful N-ACTION response.

Storage

Commit

Client

Storage

Client

Print

Client

1. Start Stereotactic Acquisition Procedure

Printer

Storage

Provider

2. Acquire Stereotactic Image on Selenia AWS

3. User Selects Stereotactic Image Will Be

Saved Locally

4. Study Completed on Selenia AWS

5. User Selects Stereotactic Images in File Manager to Be

Copied to the Previously Setup Storage DICOM Server

(Optional)

6. Send Storage Commitment if required by Storage

DICOM Server (Optional - Based on Storage DICOM

Server Configuration)

7. User Selects Print Option and Printer to print

Acquired Stereotactic Image(s) (Optional)

Figure 3.1-1: DSM Image Print/Storage Sequence Diagram

Query/Retrieve:

Storage Server

AE

Query/ Retrieve

Client AE

Query/ Retrieve

Provider

1. Query Patient

2. Receive List of Series

3. Select Series

4. Retrieve Series

5. Store Images

Figure 3.1-2: DSM Query/Retrieve Sequence Diagram

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 11 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

3.2. AE Specifications

3.2.1 DSM Client AEs

The DSM, using commercial DICOM library software, provides Standard Conformance to the following DICOM SOP Classes as an SCU:

Table 3.2.1-1 - SOP Classes For DSM Client AEs

SOP Class Name

Basic Grayscale Print Management Meta

Basic Film Session

Basic Film Box

Basic Grayscale Image Box

Printer

Digital Mammography X-Ray Image Storage – For

Presentation

SOP Class UID

1.2.840.10008.5.1.1.9

1.2.840.10008.5.1.1.1

1.2.840.10008.5.1.1.2

1.2.840.10008.5.1.1.4

1.2.840.10008.5.1.1.16

1.2.840.10008.5.1.4.1.1.1.2

Storage Commitment Push Model

Study Root Query/Retrieve Information Model – FIND

1.2.840.10008.1.20.1

1.2.840.10008.5.1.4.1.2.2.1

Study Root Query/Retrieve Information Model – MOVE 1.2.840.10008.5.1.4.1.2.2.2

Verification

1.2.840.10008.1.1

3.2.1.1 Association Policies

3.2.1.1.1 General

The DSM Client AEs are finite modules. They exist only one at a time, for the duration of a single

Association. The Print Client will attempt to establish an association when the user invokes the Print

Screen or Print Film feature. The Storage Client will attempt to establish an association when the user invokes the File Manager Copy or Auto Archive feature. The Storage Commitment Client may attempt to establish an association during Auto Archive. The Query/Retrieve Client will attempt to establish an association when the user invokes the DICOM Query feature. The Verification Client will attempt to establish an association when the user invokes DICOM Echo for one of the configured remote

DICOM servers. The maximum PDU size requested is 31K bytes.

Application Context Name

Table 3.2.1-2 - DICOM Application Context

1.2.840.10008.3.1.1.1

3.2.1.1.2 Number of Associations

Each Client AE will attempt only one association establishment at a time. Each Print, Storage,

Storage Commitment, Query/Retrieve, or Verification request results in a separate, sequential association establishment and release.

3.2.1.1.3 Asynchronous Nature

Asynchronous operations are not supported. The Print Client does however accept the asynchronous

N-EVENT-REPORT message as defined in the supported SOP Classes, provided it is sent while the association is open.

3.2.1.1.4 Implementation Identifying Information

The Implementation Class UID is: 2.16.840.1.113816.40.19991231

The Implementation Version Name is: 1.0.3

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 12 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

3.2.1.2 Association Initiation Policy

A Client AE will establish a new association each time the user invokes features that activate DICOM

Print, Storage, Storage Commitment, Query/Retrieve, or Verification. The association is released after a request has been completed.

The Print Client AE does not have a timeout while waiting for an association response from the

Remote Print SCP.

3.2.1.2.1 Modality Worklist

Refer to the Selenia Acquisition Workstation DICOM Conformance Statement for information regarding the Modality Worklist Client AE.

3.2.1.2.2 Print Screen / Print Film

3.2.1.2.2.1 Description and Sequencing of Activities

The Print Client is responsible for defining the contents of one or more films, and requesting that they be printed by the remote DICOM Print SCP. A single Film Session is defined with a number of copies and priority. Each associated Film Box is defined with format (i.e. number of images and layout), orientation, film size, and magnification. Each Basic Grayscale Image Box within a Film Box is defined with image size (rows, columns), position, bits allocated, bits stored, high bit, photometric interpretation, and the preformatted image pixel data. All of this information is transmitted to the remote DICOM Print SCP within a DICOM network association, using the Basic Grayscale Print

Management Meta SOP Class, with its associated SOP Classes and DIMSE services.

3.2.1.2.2.2 Proposed Presentation Contexts

The Print Client attempts to establish associations using the following presentation contexts:

Name

Table 3.2.1.2.2-1 - Proposed Presentation Contexts For Print Screen/Film

Presentation Context Table

Abstract Syntax

UID

Basic Grayscale

Print Management

Meta

1.2.840.10008.5.1.9

Transfer Syntax

Name List

Implicit VR,

Little Endian

Explicit VR,

Little Endian

UID List

1.2.840.10008.1.2

1.2.840.10008.1.2.1

Role Extended

Negotiation

SCU None

3.2.1.2.2.3 SOP Specific Conformance for Print SOP Classes

The general behavior of the Print Client AE during communication failure is summarized in the Table below. This behavior is common to all SOP Classes supported by the Print Client AE.

Exception

Timeout

Table 3.2.1.2.2-2 - DICOM Command Communication Failure Behavior

Association aborted by SCP

Behavior

The Print Client AE does not have a timeout while waiting for a DIMSE response from the Remote Print SCP.

DICOM Error Message will be displayed with the failed DICOM command, error code and reason. Print request is aborted.

3.2.1.2.2.4 Printer SOP Class

N-GET

Table 3.2.1.2.2-3 – Printer SOP Class: Supported DIMSE operations

Name Description

Used to retrieve the current status of the Printer.

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 13 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

Attribute Name

Printer Status

Printer Status Info

Printer Name

Manufacturer

Table 3.2.1.2.2-4 – Printer SOP Class: Supported Attributes

Tag

(2110,0010)

(2110,0020)

(2110,0030)

(0008,0070)

Supported Values

None

None

None

None

Default Value

The behavior of Print Client AE when encountering status codes in an N-GET response is summarized in the Table below.

Service

Status

Success

Warning

Failure

Table 3.2.1.2.2-5 - Printer SOP Class Response Status Handling Behavior

Further

Meaning

Success

Any

Any

Error

Code

0000

Any

Any

Behavior

Proceed to create film session if Printer Status is NORMAL.

Otherwise, a DICOM Error Message will be displayed with the failed

DICOM command, error code and reason. The print request is aborted.

A DICOM Error Message will be displayed with the failed DICOM command, error code and reason. The print request is aborted.

A DICOM Error Message will be displayed with the failed DICOM command, error code and reason. The print request is aborted.

3.2.1.2.2.5 Specific Conformance to Basic Film Session SOP Class

N-CREATE

N-DELETE

Table 3.2.1.2.2-6 - Basic Film Session SOP Class: Supported DIMSE operations

Name Description

Creates the film session

Deletes the film session

Table 3.2.1.2.2-7 - Basic Film Session SOP Class: Supported Attributes

Attribute Name

Number of Copies

Print Priority

Tag

(2000,0010)

(2000,0020)

Supported Values

1-99

HIGH

Default Value

1

HIGH

The behavior of Print Client AE when encountering status codes in an N-CREATE or N-DELETE response is summarized in the Table below.

Service

Status

Success

Warning

Failure

Table 3.2.1.2.2-8 - Film Session SOP Class Response Status Handling Behavior

Further

Meaning

Success

Memory allocation

Any

Error

Code

0000

B600

Any

Behavior

The Print SCP completed the operation successfully. Proceed to next step.

DICOM Error Message will be displayed with the failed DICOM command, error code and reason. The print request is aborted.

DICOM Error Message will be displayed with the failed DICOM command, error code and reason. The print request is aborted.

3.2.1.2.2.6 Specific Conformance to Basic Film Box SOP Class

N-CREATE

N-ACTION

Table 3.2.1.2.2-9 - Basic Film Box SOP Class: Supported DIMSE operations

Name Description

Creates the film box

Prints the film box

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 14 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

Table 3.2.1.2.2-10 - Basic Film Box SOP Class: Supported Attributes

Attribute Name

Image Display Format

Film Orientation

Film Size ID

Magnification Type

Tag

(2010,0010)

(2010,0040)

(2010,0050)

(2010,0060)

Supported Values

STANDARD\1,1

STANDARD\2,2

STANDARD\2,3

STANDARD\3,4

PORTRAIT

LANDSCAPE

8INX10IN

10INX12IN

11INX14IN

14INX14IN

14INX17IN

REPLICATE

BILINEAR

CUBIC

NONE

Default Value

None

None

None

NONE

The behavior of Print Client AE when encountering status codes in an N-CREATE or N-ACTION response is summarized in the Table below.

Service

Status

Success

Warning or

Failure

Table 3.2.1.2.2-11 - Film Box SOP Class Response Status Handling Behavior

Further

Meaning

Success

Any

Error

Code

0000

Any

Behavior

The Print SCP completed the operation successfully. Proceed to next step.

A DICOM Error Message will be displayed with the failed DICOM command, error code and reason. The print request is aborted.

3.2.1.2.2.7 Specific Conformance to Basic Grayscale Image Box SOP Class

Table 3.2.1.2.2-12 - Basic Grayscale Image Box SOP Class: Supported DIMSE operations

N-SET

Name Description

Updates an image box in a previously created film box.

Table 3.2.1.2.2-13 - Basic Grayscale Image Box SOP Class: Supported Attributes

Attribute Name

Image Position

Tag

(2020,0010)

Supported Values

1-12 depending on display format

None

Default Value

Preformatted Grayscale Image

Sequence

> Samples Per Pixel

> Photometric Interpretation

> Rows

> Columns

> Pixel Aspect Ratio

> Bits Allocated

> Bits Stored

> High Bit

> Pixel Representation

> Pixel Data

(2020,0110)

(0028,0002)

(0028,0004)

(0028,0010)

(0028,0011)

(0028,0034)

(0028,0100)

(0028,0101)

(0028,0102)

(0028,0103)

(7FE0,0010)

1

MONOCHROME2

1024

1024

1\1

8

8

7

0

1

MONOCHROME2

None

None

1\1

8

8

7

0

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 15 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

The behavior of Print Client AE when encountering status codes in an N-SET response is summarized in the Table below:

Success

Failure

Table 3.2.1.2.2-14

Basic Grayscale Image Box SOP Class N-SET Response Status Handling Behavior

Service

Status

Further

Meaning

Success

Any

Error

Code

0000

Any

Behavior

The Print SCP completed the operation successfully. Proceed to next step.

A DICOM Error Message will be displayed with the failed DICOM command, error code and reason. The print request is aborted.

3.2.1.2.3 File Manager Copy / Auto Archive

3.2.1.2.3.1 Description and Sequencing of Activities

The Storage Client as part of the File Manager Copy and Auto Archive features is responsible for defining the contents of a Digital Mammography X-Ray Image Instance and requesting that it be stored by a remote DICOM Storage SCP. One image at a time is transmitted to the remote DICOM

Storage SCP within a DICOM network Association, using the Digital Mammography X-Ray Image

Storage – For Presentation SOP Class, with its associated DIMSE services.

Level 2 (Full) conformance is expected from the DICOM storage AE.

3.2.1.2.3.2 Proposed Presentation Contexts

The Storage Client attempts to establish associations using the following presentation contexts:

Table 3.2.1.2.3-1 - Proposed Presentation Contexts For File Manager Copy / Auto Archive

Presentation Context Table

Abstract Syntax

Name

Digital

Mammography X-

Ray Image Storage

– For Presentation

UID

1.2.840.10008.5.1.4.1.1.1.2

Transfer Syntax

Name List UID List

Implicit VR,

Little Endian 1.2.840.10008.1.2

Explicit VR,

Little Endian 1.2.840.10008.1.2.1

Role Extended

Negotiation

SCU None

3.2.1.2.3.3 SOP Specific Conformance for Storage SOP Class

The behavior of DSM during communication failure is summarized in the Table below.

Exception

Timeout

Association aborted by

SCP

Table 3.2.1.2.3-2 - Storage Communication Failure Behavior

Behavior

A DICOM Error Message will be displayed with the failed DICOM command, error code and reason. Store request is aborted.

A DICOM Error Message will be displayed with the failed DICOM command, error code and reason. Store request is aborted.

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 16 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

The behavior of DSM when receiving status codes in a C-STORE response is summarized in the

Table below.

Service

Status

Success

Warning

Warning

Failure

Table 3.2.1.2.3-3 Storage C-STORE Response Status Handling Behavior

Further

Meaning

Success

Error

Code

0000

Behavior

The SCP successfully stored the SOP Instance. Proceed to next step.

Coercion of

Data

Elements

Any

B000

Any

The system has a configurable parameter that when enabled, treats a C-

STORE response with a status code of B000 as a successful operation.

Any Any

A DICOM Error Message will be displayed with the failed DICOM command, error code and reason. The store request is aborted.

A DICOM Error Message will be displayed with the failed DICOM command, error code and reason. The store request is aborted.

Instance UIDs are globally unique for all SOP instances generated by DSM. The UID root is

2.16.840.1.113816.

3.2.1.2.4 Auto Archive

3.2.1.2.4.1 Description and Sequencing of Activities

The Storage Commitment Client as part of the Auto Archive feature is responsible for defining a transaction UID and a list of Digital Mammography X-Ray Image – For Presentation SOP Class and

SOP Instance UIDs for which to request storage commitment. A single request is transmitted to a remote DICOM Storage Commitment SCP within a DICOM network association, using the Storage

Commitment Push Model SOP Class, with its associated DIMSE services. The request contains the

SOP Class and SOP Instance UIDs for all images successfully stored during Auto Archive. After successful N-ACTION, the Storage Commitment Client will wait a configurable number of minutes

(commit timeout) for the SCP to request an Association and send the N-EVENT-REPORT.

It is preferred that the remote AE of which storage commitment is requested be the same remote AE to which the images were stored.

3.2.1.2.4.2 Proposed Presentation Contexts

The Storage Commitment Client attempts to establish associations using the following presentation contexts:

Name

Storage

Commitment

Push Model

Table 3.2.1.2.4-1

Proposed Presentation Contexts For Auto Archive

Abstract Syntax

UID

Presentation Context Table

Transfer Syntax

UID List

1.2.840.10008.1.20.1

Name List

Implicit VR,

Little Endian

Explicit VR

Little Endian

1.2.840.10008.1.2

1.2.840.10008.1.2.1

Role Extended

Negotiation

SCU None

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 17 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

3.2.1.2.4.3 SOP Specific Conformance for Storage Commitment SOP Class

The behavior of DSM during communication failure is summarized in the Table below.

Exception

Timeout

Table 3.2.1.2.4-2

STORAGE COMMITMENT COMMUNICATION FAILURE BEHAVIOR

Association aborted by SCP

Behavior

A DICOM Error message will be displayed with the failed DICOM command, error code and reason. The Storage Commitment request is aborted.

A DICOM Error Message will be displayed with the failed DICOM command, error code and reason. The Storage Commitment request is aborted.

The behavior of DSM when encountering status codes in a Storage Commitment N-ACTION response is summarized in the Table below.

Table 3.2.1.2.4-3

STORAGE COMMITMENT N-ACTION RESPONSE STATUS HANDLING BEHAVIOR

Service

Status

Success

Warning

Failure

Further Meaning Error

Success

Any

Any

Code

0000

Any

Any

Behavior

Confirmation the request for Storage Commitment was successfully acknowledged by the SCP.

A DICOM Error Message will be displayed with the failed

DICOM command, error code and reason. The Storage

Commitment request is aborted.

A DICOM Error Message will be displayed with the failed

DICOM command, error code and reason. The Storage

Commitment request is aborted.

The Storage Commitment Client supports the following N-ACTION attributes:

Action

Type Name

Table 3.2.1.2.4-4 - Supported Storage Commitment N-ACTION Attributes

Action

Type ID

Attribute Tag Notes

Request

Storage

Commitment

1

Transaction UID

Referenced SOP

Sequence

> Referenced SOP Class

UID

> Referenced SOP

Instance UID

(0008,1195) DSM generates

(0008,1199)

DSM generates, one or more

Items

(0008,1150) 1.2.840.10008.5.1.4.1.1.1.2

(0008,1155)

SOP Instance UID of image to be committed

3.2.1.2.5 Query/ Retrieve Request

3.2.1.2.5.1 Description and Sequencing of Activities

The Query/Retrieve Client uses the Hierarchical Search. It is responsible for defining Study and

Series level queries (Find), followed by one or more Series level retrieve (Move) requests. All of this information is transmitted to a remove DICOM Query/Retrieve SCP within a DICOM network

Association, using the Study Root Query/Retrieve Information Model – FIND and Study Root

Query/Retrieve Information Model – MOVE SOP Classes, with their associated DIMSE services. It is intended for a DSM to retrieve only Series of images that it has stored previously to the queried SCP.

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 18 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

3.2.1.2.5.2 Proposed Presentation Contexts

The Query/Retrieve Client attempts to establish associations using the following presentation contexts:

Table 3.2.1.2.5-1 - PROPOSED PRESENTATION CONTEXTS FOR QUERY/RETRIEVE REQUEST

Presentation Context Table

Abstract Syntax

Name

Study Root

Query/Retrieve

Information Model

– FIND

Study Root

Query/Retrieve

Information Model

– MOVE

UID

1.2.840.10008.5.1.4.1.2.2

.1

1.2.840.10008.5.1.4.1.2.2

.2

Transfer Syntax

Name List

Implicit VR,

Little Endian

Explicit VR

Little Endian

Implicit VR,

Little Endian

Explicit VR

Little Endian

UID List

1.2.840.10008.1.2

1.2.840.10008.1.2.1

1.2.840.10008.1.2

1.2.840.10008.1.2.1

Role Extended

SCU

SCU

Negotiation

None

None

3.2.1.2.5.3 SOP Specific Conformance for Query/ Retrieve SOP Classes

The behavior of the DSM during communication failure is summarized in the Table below.

Table 3.2.1.2.5-2 - C-FIND AND C-MOVE COMMUNICATION FAILURE BEHAVIOR

Exception

Timeout

Association aborted by SCP

Behavior

A DICOM Error Message will be displayed with the failed DICOM command, error code and reason.

A DICOM Error Message will be displayed with the failed DICOM command, error code and reason.

The behavior of the DSM when encountering status codes in a C-FIND or C-MOVE response is summarized in the Table below.

Service

Status

Success

Pending

Warning or

Failure

Table 3.2.1.2.5-3 - C-FIND AND C-MOVE Response Status Handling Behavior

Further Meaning Error

Code

Matching or suboperations are complete

Matches are continuing

Any

0000

Behavior

successfully.

FF00, FF01 C-FIND: matching results are collected for display to user

Any

C-FIND results are displayed to the user.

Indicates all C-MOVE sub-operations completed

A DICOM Error Message will be displayed with the failed

DICOM command, error code and reason.

The Query/Retrieve Client provides standard conformance to the Study Root Information Model.

Using the hierarchical search method, a Study level query is sent, and the responses are used to send follow-up Series level queries automatically. The first 20 Series level matching responses received are displayed to the user, and the rest are discarded. For retrieve, a Series level move request is sent, containing the Study Instance UID and Series Instance UID of the series to retrieve.

For the Study level query, either Patient ID or Patient’s Name (or both) will contain a value, and Study

Date may contain a value. The user supplies Matching Key Attribute values other than Modality, either by manual entry or date range selection.

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 19 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

Table 3.2.1.2.5-4 - Supported C-FIND Matching Key Attributes (Hierarchical Search Method)

Matching Key Attribute Tag Matching Type

Study Level

Patient’s Name

Patient ID

Study Date

Study Time

Accession Number

Study ID

Study Instance UID

Series Level

Study Instance UID

Series Number

Series Instance UID

Modality

(0010,0010)

(0010,0020)

(0008,0020)

(0008,0030)

(0008,0050)

(0020,0010)

(0020,000D)

(0020,000D)

(0020,0011)

(0020,000E)

(0008,0060)

Single Value, Wild Card or Universal

Single Value, Wild Card or Universal

Single Value or Universal

Universal

Universal

Universal

Universal

Single Value (from Study level result)

Universal

Universal

Single Value = “MG”

3.2.1.2.6 DICOM Echo

3.2.1.2.6.1 Description and Sequencing of Activities

The Verification Client is responsible for performing an echo command with one of the configured remote DICOM Print, Storage or Query/Retrieve servers. A single request is transmitted to a remote

SCP within a DICOM network Association, using the Verification SOP Class, with its associated

DIMSE services.

3.2.1.2.6.2 Proposed Presentation Contexts

The Verification Client attempts to establish associations using the following presentation contexts:

Table 3.2.1.2.6-1

Proposed Presentation Contexts For DICOM ECHO

Abstract Syntax

Name

Verification SOP

Class

UID

Presentation Context Table

Transfer Syntax

1.2.840.10008.1.1

Name

Implicit VR,

Little Endian

UID

Role Extended

Negotiation

1.2.840.10008.1.2 SCU None

3.2.1.2.6.3 SOP Specific Conformance for Verification SOP Class

The behavior of DSM when encountering status codes in a C-ECHO response is summarized in the

Table below.

Service

Status

Success

Failure

Table 3.2.1.2.6-2

C-ECHO Response Status Handling Behavior

Further Meaning Error

Code

Success 0000

Behavior

Success is reported to the user.

Any Any A DICOM Error Message will be displayed with the failed

DICOM command, error code and reason.

3.2.1.3 Association Acceptance Policy

The Print, Storage, Query/Retrieve and Verification Clients do not accept Associations.

The Storage Commitment Client does accept as Association, for the Storage Commitment Push

Model SOP Class only, in order to receive the N-EVENT-REPORT notification from the Storage

Commitment SCP, containing storage commitment results.

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 20 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

3.2.1.3.1 Auto Archive – Receive Storage Commitment Response

3.2.1.3.1.1 Description and Sequencing of Activities

The Storage Commitment Client AE will accept associations in order to receive responses to a

Storage Commitment Request.

The Storage Commitment Client AE accepts a reverse role association request from a remote storage commitment AE using the Storage Commitment Push Model SOP Class. After accepting an association an N-EVENT-REPORT request is expected that provides the status of a previous request for storage commitment of one or more images. The Storage Commitment Client AE waits for a configurable period of time (commit timeout), and if an N-EVENT-REPORT request is not received, the association is released.

3.2.1.3.1.2 Accepted Presentation Contexts

Name

Storage

Commitment

Push Model

Table 3.2.1.3.1-1

Acceptable Presentation Contexts For

AUTO ARCHIVE Receive STORAGE COMMITMENT Response

Presentation Context Table

Abstract Syntax

UID

1.2.840.10008.1.20.1

Transfer Syntax

Name UID

Implicit VR,

Little Endian

Explicit VR

Little Endian

1.2.840.10008.1.2

1.2.840.10008.1.2.1

Role Extended

Negotiation

SCU None

3.2.1.3.1.3 SOP Specific Conformance for Storage Commitment SOP Class

Upon receipt of an N-EVENT-REPORT request, the commit status of the referenced images is updated, and an N-EVENT-REPORT response is sent.

The behavior of Storage Commitment Client AE when receiving Event Types within the N-EVENT-

REPORT is summarized in the Table below.

Event Type

Name

Table 3.2.1.3.1-2 - STORAGE COMMITMENT N-EVENT-REPORT Behavior

Event

Type

ID

1

Behavior

Storage

Commitment

Request

Successful

Storage

Commitment

Request

Complete –

Failures Exist

2

The archive flag of the locally stored image file is reset, to indicate the image is eligible for deletion. The system will no longer attempt to archive the image to the remote DICOM storage server.

A DICOM Error Message will be displayed with the failed DICOM command, error code and reason. The next time Auto Archive is run, it will attempt to commit the image again.

The reasons for returning specific status codes in an N-EVENT-REPORT response are summarized in the Table below.

Table 3.2.1.3.1-3 - STORAGE COMMITMENT N-EVENT-REPORT RESPONSE STATUS

Service

Status

Success

Further Meaning

Success

Error Code

0000

Reason

Storage Commitment confirmation acknowledged.

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 21 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

3.2.2 Storage Server AE

This Application Entity provides Standard Conformance to the following SOP Classes:

Table 3.2.2-1

SOP CLASSES FOR STORAGE SERVER AE

SOP Class Name

Digital Mammography X-Ray Image Storage – For

Presentation

SOP Class UID

1.2.840.10008.5.1.4.1.1.1.2

SCU

No

3.2.2.1 Association Policies

3.2.2.1.1 General

The maximum PDU size requested is 31K bytes.

The DICOM standard Application context shall be specified.

Application Context Name

Table 3.2.2-2

DICOM APPLICATION CONTEXT

1.2.840.10008.3.1.1.1

SCP

Yes

3.2.2.1.2 Number of Associations

Only one association is accepted at a time, to service storage requests from a remote DICOM

Storage SCU, such as in response to a retrieve (MOVE) request initiated by the Query/Retrieve

Client.

3.2.2.1.3 Asynchronous Nature

DSM does not support asynchronous operations (multiple outstanding transactions over a single

Association).

3.2.2.1.4 Implementation Identifying Information

Table 3.2.2-3

- DICOM IMPLEMENTATION CLASS AND VERSION FOR STORAGE SERVER

Implementation Class UID

2.16.840.1.113816.40.19991231

Implementation Version Name 1.0.3

3.2.2.2 Association Initiation Policy

The Storage Server AE does not initiate Associations.

3.2.2.3 Association Acceptance Policy

The Storage Server is always running, listening for an association request.

The Storage Server accepts an association after matching the called AE Title. All associations must use the same destination TCP/IP port number. When the Storage Server AE accepts an association, it will respond to storage requests.

3.2.2.3.1 Receive Requested Images

3.2.2.3.1.1 Description and Sequencing of Activities

The Storage Server AE is responsible for receiving and storing the images sent by a remote DICOM

Storage SCU so they are available to the user. Images are received from the remote DICOM Storage

SCU within a DICOM network association, using the Digital Mammography X-Ray Image Storage –

For Presentation SOP Class, with its associated DIMSE services. The Storage Server AE is also responsible for responding to a verification request made by the Storage SCU. The Storage Server accepts associations using the following presentation contexts:

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 22 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

Name

Verification

Digital

Mammography X-

Ray Image

Storage – For

Presentation

Table 3.2.2-4

ACCEPTABLE PRESENTATION CONTEXTS FOR

RECEIVE REQUESTED IMAGES

Presentation Context Table

Abstract Syntax

UID

1.2.840.10008.1.1

1.2.840.10008.5.1.4.1.1.1

.2

Transfer Syntax

Name UID

Implicit VR,

Little Endian

Explicit VR

Little Endian

Implicit VR,

Little Endian

1.2.840.10008.1.2

1.2.840.10008.1.2.1

1.2.840.10008.1.2

Explicit VR

Little Endian

1.2.840.10008.1.2.1

Role Extended

Negotiation

SCP None

SCP None

3.2.2.3.1.3 SOP Specific Conformance for Storage SOP Class

The Storage Server AE provides Level 2 conformance to the Storage Service Class.

The Storage Server AE will behave as described in the Table below when generating the C-STORE response.

Table 3.2.2-5 - STORAGE C-STORE RESPONSE STATUS

Service

Status

Success

Further Meaning Error Code

0000

Reason

Failure

Failure

Success

Refused: Out of

Resources

Missing or Invalid

Attribute

A700

0106

Image is stored locally and made available to the user.

There is not enough local storage capacity to store the image.

The Pixel Data was not read successfully.

3.3. Network Interfaces

3.3.1 Physical Network Interface

The DSM is tested and supports using 10/100/1000-BaseT Ethernet media.

3.3.2 Additional Protocols

None.

3.4. Configuration

The Print Client configuration information is stored in a binary data file accessible to the DSM. All configuration parameters may be viewed and/or modified from the DICOM Print Film Setup dialog, which is accessed via the DICOM Print Setup menu item on the DSM Setup Menu.

The Storage Client, Storage Commitment Client, Query/Retrieve Client, and Storage Server configuration information is stored in a binary data file accessible to the DSM. All configuration parameters may be viewed and/or modified from the DICOM Server Setup dialog, which is accessed via the DICOM Server Setup menu item on the DSM Setup Menu.

Note: Refer to the Selenia Acquisition Workstation DICOM Conformance for Modality Worklist configuration information.

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 23 of 32

3.4.1 AE Title/ Presentation Address Mapping

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

3.4.1.1 Local AE Titles

Application Entity

Storage Client

Print Client

Storage Commitment Client

Query/Retrieve Client

Storage Server

Table 3.4-1

AE TITLE CONFIGURATION TABLE

Default AE Title

DSM_MG_SCU

DSM_MG_SCU

DSM_MG_SCU

DSM_MG_SCU

DSM_MG_SCP

N/A

N/A

104

N/A

104

Default TCP/IP Port

The SCU AE Title may be modified in the DICOM Print Film Setup dialog. This AE Title applies to all

Client AEs.

The Storage Server AE Title and port number may be modified in the DICOM Server Setup dialog.

3.4.1.2 Remote AE Title/ Presentation Address Mapping

Up to three Remote Storage SCPs can be configured as well as up to three Remote Print SCPs.

3.4.1.2.1 Remote Modality Worklist SCP

Note: Refer to the Selenia Acquisition Workstation DICOM Conformance for Modality Worklist information.

3.4.1.2.2 Remote Print SCP

The following are the configurable items for each Remote Print SCP in the DICOM Print Film Setup dialog:

Parameter

Application Entity Title

(Called AE)

Remote Host/IP Address

Remote Port Number

Printer Name

Film Size

Magnification Type

Table 3.4-2 – Remote Print SCP Configuration

Value Description

The DSM supports multiple AEs acting as Remote Print SCP.

IP address and hostname, one per Remote Print AE.

Port number to use when establishing TCP/IP connection, one per Remote

Print AE.

User defined printer name that appears on the Print Screen/Print Film destination list

8INX10IN, 11INX14IN, 14INX14IN, 14INX17IN

None, Replicate, Bilinear, Cubic

3.4.1.2.3 Remote Storage, Storage Commitment, Query/ Retrieve SCP

The following are the configurable items for each Remote Storage SCP in the DICOM Server Setup dialog:

Parameter

DICOM Server AE Title

DICOM Server Hostname

TCP/IP Port Number

DICOM Server Name

Table 3.4-3 – Remote Storage SCP Configuration

Value Description

The DSM supports up to three AEs acting as Remote Storage, Storage

Commitment and/or Query/Retrieve SCP.

IP address and hostname, one per Remote Storage AE.

Port number to use when establishing TCP/IP connection, one per Remote

Storage AE.

User defined server name that appears in the File Manager Copy destination list, Auto Archive Setup destination list, and/or Query/Retrieve destination list

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 24 of 32

Parameter

DICOM Request Timeout

DICOM Commit Timeout

Mammography Storage

Request Storage

Commitment for Archive

Query/Retrieve

Treat Coercion Warnings as Success

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

Value Description

Default = 30 seconds, range = 1-65535 seconds, controls time to wait for association response and C-STORE, N-ACTION, C-FIND and C-MOVE responses.

Default = 60 minutes, range = 1-65535 minutes, controls time to wait for N-

EVENT-REPORT request that confirms storage commitment.

Enable or Disable

Enable or Disable

Enable or Disable

Enable or Disable

3.4.2 Configurable Parameters for Local AEs

The following are the configurable items for the local AEs:

Table 3.4-4 – CONFIGURATION PARAMETERS

Parameter Configurable

(Yes/No)

Storage Client

Study Description Format (if value not provided by Selenia)

Yes

Print Client

Film Format (1 x 1, 2 x 2, 3 x 2, 4 x 3)

Print Resolution 1024x1024

Institution Name (applies to Created SOP Instances also)

Institution Address (applies to Created SOP Instances also)

Query/Retrieve Client

Allow Partial Patient ID’s

Allow Partial Patient Name

Yes

No

Yes

Yes

Yes

Yes

Default Value

Digital Spot

Compression: value of Study ID

High (1024)

No

Yes

4.0 Media Interchange

Not Applicable

5.0 Support of Character Sets

The DSM supports the following character sets for the values of Data Elements with a VR of SH, LO,

ST, PN or LT:

• Latin alphabet No. 1 (ISO-IR 100 & ISO-IR 6)

6.0 Security

The DSM does not support any specific DICOM security measures.

It is assumed that the DSM is used within a secured environment. It is assumed that a secured environment includes at a minimum:

• Firewall or router protections to ensure that only approved external hosts have network access to the DSM.

• Firewall or router protections to ensure that the DSM only has network access to approved external hosts and services.

• Any communication with external hosts and services outside the locally secured environment use appropriate secure network channels (e.g. such as a Virtual Private Network (VPN)).

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 25 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

• Other network security procedures such as automated intrusion detection may be appropriate in some environments. Additional security features may be established by the local security policy and are beyond the scope of this conformance statement.

7.0 Annexes

7.1. IOD Contents

7.1.1 Created SOP Instance(s)

Abbreviations for Presence of Value:

ALWAYS Attribute always present with a value

ANAP Attribute not always present

EMPTY Attribute is sent without a value (zero length)

VNAP Attribute value not always present (zero length if no value is present)

Abbreviations for Source:

AUTO Attribute value is generated automatically

CONFIG Attribute value source is a configurable parameter

MWL

USER

Attribute value is the same as the value received using Modality Worklist

Attribute value source is from User input

7.1.1.1 Digital Mammography X-Ray Image IOD

The default attributes that are contained in created SOP Instances are listed in the following tables.

The attributes contained in created SOP Instances are configurable.

IE

Patient

Study

Series

Image

Table 7.1-1

IOD OF CREATED DIGITAL MAMMOGRAPHY X-RAY IMAGE SOP INSTANCES

Module

Patient

Reference

Table 7.1-2

Presence of Module

ALWAYS

General Study

Table 7.1-3 ALWAYS

General Series

Table 7.1-4 ALWAYS

DX Series

Table 7.1-6 ALWAYS

Mammography Series

Table 7.1-6 ALWAYS

Equipment General Equipment

General Image

Table 7.1-5

Table 7.1-7

ALWAYS

ALWAYS

Image Pixel

Table 7.1-8 ALWAYS

DX Anatomy Imaged

Table 7.1-9 ALWAYS

DX Image

Table 7.1-10 ALWAYS

DX Detector

Table 7.1-11 ALWAYS

DX Positioning

Table 7.1-12 ALWAYS

X-Ray Acquisition Dose

Table 7.1-13 ALWAYS

X-Ray Generation

Table 7.1-14 ALWAYS

X-Ray Filtration

Table 7.1-15 ALWAYS

X-Ray Grid

Table 7.1-16 ALWAYS

Mammography Image

Table 7.1-17 ALWAYS

Image Histogram

Table 7.1-18 ALWAYS

Acquisition Context

Table 7.1-19 ALWAYS

SOP Common

Table 7.1-20 ALWAYS

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 26 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

7.1.1.2 Common Modules

Attribute Name

Patient’s Name

Patient ID

Issuer of Patient ID

Patient’s Birth Date

Patient’s Sex

Other Patient IDs

Table 7.1-3 General Study Module of Created SOP Instances

Attribute Name Tag Value

Study Description

Study Instance UID

Study ID

Patient’s Age

Table 7.1-2 Patient Module of Created SOP Instances

Study Date

Study Time

Accession Number

Referring Physician’s Name

Tag Value

(0010,0010) From Selenia

(0010,0020) From Selenia

(0010,0021) From Selenia

(0010,0030) From Selenia

(0010,0040) From Selenia

(0010,1000) From Selenia

Presence of

Value

ALWAYS

ALWAYS

ANAP

VNAP

VNAP

ANAP

Presence of

Value

(0008,0020)

(0008,0030)

(0008,0050)

(0008,0090)

ALWAYS

ALWAYS

VNAP

VNAP

(0008,1030) mapped from one of two attributes, or a single entered value applied to all studies

ALWAYS

(0020,000D) Unique value from Selenia ALWAYS

(0020,0010) ALWAYS

(0010,1010)

From Selenia

From Selenia

From Selenia

From Selenia

From Selenia or if not provided:

Digital Spot Compression: value of Study ID or can be

From Selenia

Calculated from Patient’s

Birth Date (0010,0030) and

Study Date (0008,0020) if changed by user; otherwise use value from Selenia

ANAP

Attribute Name

Modality

Table 7.1-4 General Series Module of Created SOP Instances

Tag

(0008,0060)

Value Presence of

Value

Series Instance UID

Series Number

Series Date

Series Time

Series Description

Operators’ Name

Body Part Examined

(0020,000E)

(0020,0011)

(0008,0021)

(0008,0031)

(0008,103E)

(0008,1070)

(0018,0015)

See Table 7.1-6

Unique value from Selenia, each image in a different

Series

From Selenia

From Selenia

From Selenia

From Selenia

From Selenia

BREAST

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

Source

AUTO/USER

AUTO/USER

AUTO

AUTO/USER

AUTO/USER

AUTO

Source

AUTO

AUTO

AUTO/USER

AUTO/USER

AUTO/CONFIG

AUTO

AUTO/USER

AUTO

Source

AUTO

AUTO

AUTO

AUTO

AUTO

AUTO/USER

AUTO

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 27 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

Table 7.1-5 General Equipment Module of Created SOP Instances

Attribute Name

Manufacturer

Institution Name

Institution Address

Station Name

Manufacturer's Model Name

Device Serial Number

Software Versions

Tag

(0008,0070)

(0008,0080)

(0008,0081)

(0008,1010)

(0008,1090)

(0018,1000)

(0018,1020)

Value

HOLOGIC, Inc. (default)

Hostname

Digital StereoLoc II

From Selenia

Presence of

Value

ALWAYS

ANAP

ANAP

ALWAYS

ALWAYS

ALWAYS

ALWAYS

Source

CONFIG

CONFIG

CONFIG

AUTO

AUTO

AUTO

AUTO

7.1.1.3 Digital Mammography X-Ray Image Modules

Table 7.1-6 DX Series / Mammography Series Module of Created SOP Instances

Attribute Name

Modality

Presentation Intent Type

Tag

(0008,0060)

(0008,0068)

Value

MG

FOR PRESENTATION

Presence of

Value

ALWAYS

ALWAYS

Source

AUTO

AUTO

Rows

Columns

Bits Allocated

Bits Stored

High Bit

Pixel Representation

Pixel Data

Table 7.1-7 General Image Module of Created SOP Instances

Attribute Name

Instance Number

Patient Orientation

Content Date

Content Time

Image Type

Image Comments

Acquisition Date

Acquisition Time

Burned in Annotation

Lossy Image Compression

Value

From Selenia

See Table 7.1-10

From Selenia

From Selenia

See Table 7.1-10

From Selenia

From Selenia

See Table 7.1-10

See Table 7.1-10

Presence of

Value

ALWAYS

ALWAYS

ALWAYS

ANAP

ALWAYS

ALWAYS

Table 7.1-8 Image Pixel Module of Created SOP Instances

Attribute Name

Samples per Pixel

Photometric Interpretation

Tag

(0020,0013)

(0020,0020)

(0008,0023)

(0008,0033)

(0008,0008)

(0020,4000)

(0008,0022)

(0008,0032)

(0028,0301)

(0028,2110)

Tag

(0028,0002)

(0028,0004)

(0028,0010)

(0028,0011)

(0028,0100)

(0028,0101)

(0028,0102)

(0028,0103)

(7FE0,0010)

Value

1

MONOCHROME2 (default) or MONOCHROME1 (user inverted)

1024

1024

16

12

11

0

Presence of

Value

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

Source

AUTO

AUTO

AUTO

USER

AUTO

AUTO

Source

AUTO

AUTO/USER

AUTO

AUTO

AUTO

AUTO

AUTO

AUTO

AUTO

Table 7.1-9 DX Anatomy Imaged Module of Created SOP Instances

Attribute Name Tag Value Presence of

Value

The supported attributes are listed in Table 7.1-17

Source

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 28 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

Table 7.1-10 DX Image Module of Created SOP Instances

Attribute Name Tag

Image Type

Samples per Pixel

Photometric Interpretation

Bits Allocated

Bits Stored

High Bit

Pixel Representation

Pixel Intensity Relationship

(0008,0008)

(0028,0002)

(0028,0004)

(0028,0100)

(0028,0101)

(0028,0102)

(0028,0103)

(0028,1040)

Pixel Intensity Relationship Sign (0028,1041)

Rescale Intercept

Rescale Slope

Rescale Type

(0028,1052)

(0028,1053)

(0028,1054)

Presentation LUT Shape

Lossy Image Compression

Patient Orientation

Burned in Annotation

Window Center

Window Width

(2050,0020)

(0028,2110)

(0020,0020)

(0028,0301)

(0028,1050)

(0028,1051)

Value

See Table 7.1-17

See Table 7.1-8

See Table 7.1-8

See Table 7.1-8

See Table 7.1-8

See Table 7.1-8

See Table 7.1-8

LIN

1

0

1

US

IDENTITY (default) or

INVERSE (user inverted)

00

R\P for chest wall on bottom

L\A for chest wall on top

NO

1 or 2 values – Current and

Saved (if different)

1 or 2 values – Current and

Saved (if different)

Presence of

Value

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

Detector Type

Detector ID

Date of Last Detector

Calibration

Time of Last Detector

Calibration

Imager Pixel Spacing

Table 7.1-11 DX Detector Module of Created SOP Instances

Attribute Name Tag

(0018,7004)

(0018,700A)

(0018,700C)

(0018,700E)

Value

DIRECT

From Selenia

From Selenia

From Selenia

Presence of

Value

ALWAYS

ALWAYS

ALWAYS

ALWAYS

Pixel Spacing

(0018,1164)

(0028,0030)

(0054,0220)

0.070\0.070

Imager Pixel Spacing /

ERMF

See Table 7.1-17

ALWAYS

ALWAYS

Table 7.1-12 DX Positioning Module of Created SOP Instances

Attribute Name Tag Value Presence of

Value

View Code Sequence

Estimated Radiographic

Magnification Factor

Positioner Type

Positioner Primary Angle

Body Part Thickness

(0018,1114)

(0018,1508)

(0018,1510)

(0018,11A0)

1.173 (default)

See Table 7.1-17

See Table 7.1-17

From DSM

ALWAYS

ALWAYS

Source

AUTO

AUTO

AUTO

AUTO

AUTO

AUTO/USER

AUTO

AUTO

AUTO

AUTO

AUTO

Source

AUTO

AUTO

AUTO

AUTO

AUTO

AUTO

Source

CONFIG

AUTO

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 29 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

Table 7.1-13 X-Ray Acquisition Dose Module of Created SOP Instances

Attribute Name Tag Value Presence of

Value

The other supported attributes are listed in Tables 7.1-12, 7.1-14, 7.1-17

Entrance Dose

Entrance Dose in mGy

Organ Dose

(0040,0302)

(0040,8302)

(0040,0316)

Less than 1 dGy

From DSM

From Selenia

EMPTY

ALWAYS

ALWAYS

Table 7.1-14 X-Ray Generation Module of Created SOP Instances

Source

AUTO

AUTO

AUTO

Attribute Name

KVP

X-Ray Tube Current

Exposure Time

Exposure

Exposure in uAs

Exposure Control Mode

Exposure Control Mode

Description

Focal Spot

Tag

(0018,0060)

(0018,1151)

(0018,1150)

(0018,1152)

(0018,1153)

(0018,7060)

(0018,7062)

(0018,1190)

Value

From Selenia

From Selenia

From Selenia

From Selenia

From Selenia

From Selenia (e.g.

MANUAL)

From Selenia (e.g.

MANUAL)

From Selenia: 0.3

From Selenia (e.g.

TUNGSTEN)

Presence of

Value

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

Source

AUTO

AUTO

AUTO

AUTO

AUTO

AUTO

AUTO

AUTO

AUTO

Anode Target Material (0018,1191) ALWAYS

Table 7.1-15 X-Ray Filtration Module of Created SOP Instances

Attribute Name Tag Value Presence of

Value

Filter Material (0018,7050)

From Selenia (e.g.

RHODIUM)

ALWAYS

Table 7.1-16 X-Ray Grid Module of Created SOP Instances

Attribute Name Tag Value

Grid

Image Type

Positioner Type

Positioner Primary Angle

Image Laterality

Organ Exposed

Breast Implant Present

(0008,0008)

(0018,1508)

(0018,1510)

(0020,0062)

(0040,0318)

(0028,1300)

Presence of

Value

(0018,1166)

From Selenia:

HTC_IN or HTC_OUT

ALWAYS

Table 7.1-17 Mammography Image Module of Created SOP Instances

Attribute Name Tag Value Presence of

Value

ORIGINAL\PRIMARY\valu e 3, where value 3 is one of

STEREO_SCOUT

STEREO_MINUS

STEREO_PLUS

PREFIRE_MINUS

PREFIRE_PLUS

POSTFIRE_MINUS

ALWAYS

POSTFIRE_PLUS

POSTBIOPSY_MINUS

POSTBIOPSY_PLUS

POSTBIOPSY blank (for Specimen)

MAMMOGRAPHIC

From Selenia

From Selenia: R, L

BREAST

ALWAYS

ALWAYS

ALWAYS

ALWAYS

From Selenia: YES or NO ALWAYS

Source

AUTO

Source

AUTO

Source

AUTO/USER

AUTO

AUTO

AUTO/USER

AUTO

AUTO

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 30 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

Attribute Name

View Code Sequence

> Code Value (0008,0100)

> Coding Scheme Designator (0008,0102)

> Code Meaning (0008,0104)

>View Modifier Code Sequence (0054,0222)

>> Code Value (0008,0100)

>> Coding Scheme Designator (0008,0102)

>> Code Meaning (0008,0104)

Attribute Name

Tag

Partial View

Anatomic Region Sequence

(0028,1350)

(0008,2218)

> Code Value (0008,0100)

> Coding Scheme Designator (0008,0102)

> Code Meaning (0008,0104)

(0054,0220)

Tag

Value

NO

One Item

T-04000

SNM3 or SRT (default)

Breast

One Item representing CC,

FB, MLO, LMO, LM, ML

SIO or Specimen

<Default = CC, user must edit>

SNM3 or SRT (default)

One or two Items. Spot

Compression is always present. A second Item is present only when

Tangential is performed.

R-102D7

SNM3 or SRT (default)

Spot Compression

Presence of

Value

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

Table 7.1-18 Image Histogram Module of Created SOP Instances

Value Presence of

Value

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

ALWAYS

Histogram Sequence

>Histogram Number of Bins

>Histogram First Bin Value

>Histogram Last Bin Value

>Histogram Bin Width

>Histogram Explanation

>Histogram Data

(0060,3000)

(0060,3002)

(0060,3004)

(0060,3006)

(0060,3008)

(0060,3010)

(0060,3020)

One Item

Table 7.1-19 Acquisition Context Module of Created SOP Instances

Attribute Name Tag Value

Acquisition Context Sequence (0040,0555)

Presence of

Value

EMPTY

Table 7.1-20 SOP Common Module of Created SOP Instances

Attribute Name Tag Value

Specific Character Set

SOP Class UID

SOP Instance UID

Instance Number

(0008,0005)

(0008,0016)

(0008,0018)

(0020,0013)

Presence of

Value

ISO_IR 100

UID for Digital

Mammography X-Ray

Image Storage – For

ALWAYS

ALWAYS

Presentation

Unique value from Selenia ALWAYS

See Table 7.1-7

Source

AUTO

AUTO

AUTO

CONFIG

AUTO

USER

AUTO

CONFIG

AUTO

AUTO

AUTO

CONFIG

AUTO

Source

AUTO

AUTO

AUTO

AUTO

AUTO

AUTO

AUTO

Source

AUTO

Source

AUTO

AUTO

AUTO

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 31 of 32

Selenia Digital StereoLoc II DSM v4.3

DICOM Conformance Statement

7.1.2 Usage of Attributes from Received IODs

The only requirement for SOP Instance acceptance is that the Pixel Data is read successfully.

7.1.3 Attribute Mapping

Refer to the Selenia Acquisition Workstation DICOM conformance statement for information regarding DICOM Modality Worklist attribute mapping for the DSM.

7.1.4 Coerced/ Modified Attributes

Any attribute whose source is USER can be modified by the user, including certain attribute values that were obtained via Modality Worklist. See Section 7.1.1 Created SOP Instance(s) for attributes with a source of USER. The Storage Client incorporates user modified attribute values into created

SOP Instances.

If a SOP Instance is received with missing or empty Patient ID (0010,0020) or Accession Number

(0008,0050), the Storage Server assigns a generic value to the attribute so that the SOP Instance can be accepted. If Study Date (0008,0020) or Content Time (0008,0033) is missing or empty, the

Storage Server assigns the current date or time, respectively.

7.2. Data Dictionary of Private Attributes

Created SOP Instances may contain a Private Group (0071) labeled LORAD - A Hologic Company, containing proprietary image characteristics relating to target, target coordinates, reference and scout information.

7.3. Coded Terminology and Templates

Not Applicable

7.4. Grayscale Image Consistency

The DSM monitor used to perform stereotactic procedures is calibrated according to the DICOM

Grayscale Standard Display Function (GSDF).

7.5. Standard Extended/ Specialized/ Private SOP Classes

Not Applicable

7.6. Private Transfer Syntaxes

None

MAN-02018 Revision 004 Hologic, Inc

35 Crosby Drive, Bedford, MA 01730-1401 (781) 999-7300

Page 32 of 32

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals