Omron BP710 is a home blood pressure monitor that uses the oscillometric method to measure blood pressure and pulse rate. It features an IntelliSense technology that automatically adjusts the cuff pressure for a more comfortable and accurate measurement. The device also includes an irregular heartbeat detector and a hypertension indicator to alert users of potential health issues. The BP710 also has a memory function to store up to 30 readings for easy tracking and comparison over time. This compact and user-friendly design makes it an ideal choice for home blood pressure monitoring.
Hirdetés
Hirdetés
INSTRUCTION MANUAL
Automatic
Blood Pressure Monitor
Model BP710
Before using the Monitor
Introduction ......................................................................................................... 3
Safety Information ................................................................................................ 4
Operating the Device .......................................................................... 4
Care and Maintenance ......................................................................... 5
Operating Instructions
Before Taking a Measurement .............................................................................. 6
Know Your Unit ................................................................................................... 7
Unit Display ......................................................................................................... 8
Display Symbols ................................................................................................... 9
Irregular Heartbeat Symbol .................................................................. 9
Heartbeat Symbol/Hypertension Indicator ......................................... 9
Battery Installation ............................................................................................. 11
Applying the Arm Cuff ....................................................................................... 12
Applying the Cuff on the Left Arm .................................................... 12
Applying the Cuff on the Right Arm ................................................. 13
Taking a Measurement ....................................................................................... 14
Special Conditions …... ..................................................................... 15
Using the Memory Function ............................................................................... 16
To Delete All the Values Stored in Memory ..................................... 16
Care and Maintenance
............................................................................................... 17
Error Indicators ................................................................................................... 18
Troubleshooting Tips ........................................................................................... 19
FCC Statement .................................................................................................... 20
Limited Warranties ............................................................................................. 21
6SHFL¿FDWLRQV ..................................................................................................... 22
2
INTRODUCTION
Thank you for purchasing the OMRON
®
BP710 IntelliSense
®
Automatic Blood
Pressure Monitor.
Fill in for future reference.
DATE PURCHASED:__________________
SERIAL NUMBER:____________________
6WDSOH\RXUSXUFKDVHUHFHLSWKHUH
5HJLVWHU\RXUSURGXFWRQOLQHDW
ZZZUHJLVWHURPURQFRP
Your new digital blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood’s movement through your brachial artery and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope so the monitor is simple to use.
The BP710 comes with the following components:
0RQLWRU6WDQGDUG$GXOW$UP&XII
,QVWUXFWLRQ0DQXDOSULQWHGLQ(QJOLVKDQG6SDQLVK4XLFN6WDUW*XLGH
The BP710 IntelliSense ® $XWRPDWLF,QÀDWLRQ%ORRG3UHVVXUH0RQLWRULVLQWHQGHGIRU home use.
SAVE THESE INSTRUCTIONS
3
SAFETY INFORMATION
To assure the correct use of the product, basic safety measures should always be followed including the warnings and cautions listed in this instruction manual.
SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL
WARNING
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.
OPERATING THE DEVICE
&RQWDFW\RXUSK\VLFLDQIRUVSHFL¿FLQIRUPDWLRQDERXW\RXUEORRGSUHVVXUH
Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician or licensed healthcare provider.
The monitor is not intended to be a diagnostic device.
,IEDWWHU\ÀXLGVKRXOGJHWLQ\RXUH\HVLPPHGLDWHO\ULQVHZLWKSOHQW\RIFOHDQ water. Contact a physician immediately.
Read all of the information in the instruction manual and any other literature in the box before operating the unit.
This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on infants or persons who cannot express their intentions.
Read the Special Conditions section of the instruction manual if your systolic
SUHVVXUHLVNQRZQWREHPRUHWKDQPP+J,QÀDWLQJWRDKLJKHUSUHVVXUHWKDQ necessary may result in bruising where the cuff is applied.
Operate the device only as intended. Do not use the device for any other purpose.
Dispose of the device, components and optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution.
Do not use a cellular phone near the device. It may result in an operational failure.
4
SAFETY INFORMATION
Use only Omron authorized parts and accessories. Parts and accessories not approved for use with the device may damage the unit.
Use only 1.5 V “AAA” alkaline batteries with this device. Do not use other types of batteries. This may damage the unit.
CARE AND MAINTENANCE
'RQRWVXEMHFWWKHPRQLWRUWRVWURQJVKRFNVVXFKDVGURSSLQJWKHXQLWRQWKHÀRRU
Do not submerge the device or any of the components in water.
Store the device and the components in a clean, safe location.
&KDQJHVRUPRGL¿FDWLRQQRWDSSURYHGE\WKHPDQXIDFWXUHUZLOOYRLGWKHXVHU warranty. Do not disassemble or attempt to repair the unit or components.
5
BEFORE TAKING A MEASUREMENT
To ensure a reliable reading follow these recommendations:
1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes before taking a measurement. Rest for at least 15 minutes before taking the measurement.
2. Stress raises blood pressure. Avoid taking measurements during stressful times.
3. Measurements should be taken in a quiet place.
5HPRYHWLJKW¿WWLQJFORWKLQJIURP\RXUOHIWDUP
6LWRQDFKDLUZLWK\RXUIHHWÀDWRQWKHÀRRU5HVW\RXUOHIWDUPRQWDEOHVRWKDWWKH cuff is at the same level as your heart.
6. Remain still and do not talk during the measurement.
7. Wait 2-3 minutes between measurements. The wait time allows the arteries to return to the condition prior to taking the blood pressure measurement. You may need to increase the wait time depending on your individual physiological characteristics.
8. Keep a record of your blood pressure and pulse readings for your physician. A single measurement does not provide an accurate indication of your true blood pressure. You need to take and record several readings over a period of time. Try to measure your blood pressure at the same time each day for consistency.
6
Main Unit:
KNOW YOUR UNIT
Display
Battery compartment
START/STOP button
Memory (M) button
Air jack
Components:
Standard Adult Arm Cuff
Air tube
Air plug
INSTRUCTION MANUAL
Model BP710
INTRODUCTION
Thank you for purchasing the OMRON® BP710 Automatic Blood Pressure Monitor.
Fill in for future reference.
DATE PURCHASED: __________________
SERIAL NUMBER: ____________________
Instruction
Manual
Quick Start
Guide
7
Systolic blood pressure
Diastolic blood pressure
Memory symbol
(Displayed when viewing values stored in memory)
Irregular Heartbeat symbol
UNIT DISPLAY
Heartbeat symbol
Pulse display
Deflation symbol
Battery low indicator
8
DISPLAY SYMBOLS
IRREGULAR HEARTBEAT SYMBOL ( )
When the monitor detects an irregular rhythm two or more times during the measurement, the Irregular Heartbeat Symbol ( ) will appear on the display with the measurement values.
$QLUUHJXODUKHDUWEHDWUK\WKPLVGH¿QHGDVD
Normal Heartbeat
rhythm that is more than 25% slower or 25% faster from the average rhythm detected while the monitor is measuring the systolic and diastolic
blood pressure.
Pulse
Blood pressure
If the irregular heartbeat symbol ( )
Irregular Heartbeat
Short Long
displays with your measurement results,
we recommend you consult your
Pulse
physician. Follow the directions of your
doctor.
Blood pressure
HEARTBEAT SYMBOL/HYPERTENSION INDICATOR ( )
TAKING A MEASUREMENT
The Heartbeat Symbol ( ÀDVKHVRQWKHGLVSOD\GXULQJWKHPHDVXUHPHQW7KH
Heartbeat Symbol ( ÀDVKHVDWHYHU\KHDUWEHDW
When the measurement is complete, the Heartbeat Symbol ( ÀDVKHVRQWKH display with your blood pressure and pulse rate if the reading is above 135 for the
Systolic Blood Pressure value and/or above 85 for the Diastolic Blood Pressure value.
USING THE MEMORY FUNCTION
The Heartbeat Symbol ( ÀDVKHVZKHQWKHEORRGSUHVVXUHUHDGLQJVWRUHGLQWKH memory is above 135 for the Systolic Blood Pressure value and/or above 85 for the
Diastolic Blood Pressure value.
9
DISPLAY SYMBOLS
The American Heart Association recommends the following guideline for upper limit of normal for home blood pressure readings:
AHA Home Guideline for Upper Limit of Normal BP
Systolic Blood Pressure 135 mmHg
Diastolic Blood Pressure 85 mmHg
This is a general guideline because blood pressure depends on age and morbidity.
The AHA recommends a lower home blood pressure goal for certain patients, including diabetic patients, pregnant women, and patients with renal failure 1 .
WARNING
The monitor is not intended to be a diagnostic device.
WARNING
&RQWDFW\RXUSK\VLFLDQIRUVSHFL¿FLQIRUPDWLRQDERXW\RXUEORRGSUHVVXUH
Self-diagnosis and treatment using measured results may be dangerous.
Follow the instructions of your physician or licensed healthcare provider.
Omron Healthcare recommends you contact your physician if the Heartbeat Symbol
( ÀDVKHVRQWKHGLVSOD\ZLWK\RXUPHDVXUHPHQWYDOXHV
1 $PHULFDQ+HDUW$VVRFLDWLRQ6FLHQWL¿F6WDWHPHQW
10
BATTERY INSTALLATION
CAUTION
Use only 1.5V “AAA” alkaline batteries with this device.
3UHVVWKHŸLQGLFDWRURQWKHEDWWHU\FRYHUDQGVOLGH the cover off in the direction of the arrow.
2. Install 4 “AAA” size batteries so the + (positive) and -
(negative) polarities match the polarities of the battery compartment as indicated.
NOTE:
4 “AAA” size batteries are not included in this unit.
3. Replace the battery cover.
BATTERY REPLACEMENT
Low Battery Indicator
When the Low Battery Indicator appears on the display, turn the monitor off and remove all the batteries. Replace with four new batteries at the same time. Long-life alkaline batteries are recommended.
WARNING
,IEDWWHU\ÀXLGVKRXOGJHWLQ\RXUH\HVLPPHGLDWHO\ULQVHZLWKSOHQW\RI clean water. Contact a physician immediately.
CAUTION
Dispose of the device, components and optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution.
11
APPLYING THE ARM CUFF
CAUTION
Read all of the information in the instruction manual and any other literature in the box before operating the unit.
CAUTION
This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on infants or persons who cannot express their intentions.
APPLYING THE CUFF ON THE LEFT ARM
1. Make sure the air plug is securely inserted in the main unit.
5HPRYHWLJKW¿WWLQJFORWKLQJIURP\RXUXSSHUDUP
3. Apply the cuff to your left upper arm so
the arrow is centered on the inside of your
DUPDQGDOLJQHGZLWK\RXUPLGGOH¿QJHU
The air tube runs down the inside of your arm. The bottom of the cuff should be
approximately 1/2" above your elbow.
Left Arm
1/2 inch (1 - 2 cm)
Arrow
Air Tube
12
APPLYING THE ARM CUFF
:UDSWKHFXII¿UPO\LQSODFHDURXQG\RXU
arm using the cloth fastening.
APPLYING THE CUFF ON THE RIGHT ARM
When taking a measurement using the right arm, use this instruction for Step 3.
Right Arm
Apply the cuff to your right upper arm so the arrow and the air tube runs along the inside
of your arm. The bottom of the cuff should be
approximately 1/2" above your elbow.
Arrow
NOTE:
%HFDUHIXOQRWWRUHVW\RXUDUPRQWKHDLUWXEH7KLVZLOOUHVWULFWWKHÀRZRIDLU to the cuff.
13
TAKING A MEASUREMENT
6LWRQDFKDLUZLWK\RXUIHHWÀDWRQWKHÀRRU
Place your arm on a table so the cuff is level
with your heart.
Keep still and do not talk during measurement.
2. Press the START/STOP button.
All symbols appear on the display.
7KHFXIIVWDUWVWRLQÀDWHDXWRPDWLFDOO\$VWKHFXIILQÀDWHVWKHPRQLWRU
DXWRPDWLFDOO\GHWHUPLQHV\RXULGHDOLQÀDWLRQOHYHO
Remain still and do not move your arm until the entire measurement process is completed.
NOTE:
7RVWRSWKHLQÀDWLRQRUPHDVXUHPHQWSXVKWKH67$576723EXWWRQ7KH
PRQLWRUZLOOVWRSLQÀDWLQJVWDUWGHÀDWLQJDQGZLOOWXUQRII
,QÀDWLRQVWRSVDXWRPDWLFDOO\DQGWKHPHDVXUHPHQWLVVWDUWHG
$VWKHFXIIGHÀDWHVGHFUHDVLQJQXPEHUVDSSHDURQWKHGLVSOD\DQGWKH+HDUWEHDW
Symbol ( ÀDVKHV
4. When the measurement is complete, the arm cuff completely
GHÀDWHV<RXUEORRGSUHVVXUHDQGSXOVHUDWHDUHGLVSOD\HG
14
TAKING A MEASUREMENT
5. Press the START/STOP button to turn the monitor off.
NOTE:
7KHPRQLWRUZLOODXWRPDWLFDOO\WXUQRIIDIWHU¿YHPLQXWHV
SPECIAL CONDITIONS
CAUTION
,QÀDWLQJWRDKLJKHUSUHVVXUHWKDQQHFHVVDU\PD\UHVXOWLQEUXLVLQJZKHUHWKHFXIILV applied.
If your systolic pressure is known to be more than 220 mmHg, press and hold the
67$576723EXWWRQXQWLOWKHPRQLWRULQÀDWHVWRPP+JKLJKHUWKDQ\RXU expected systolic pressure.
NOTES:
'RQRWDSSO\PRUHSUHVVXUHWKDQQHFHVVDU\
7KHPRQLWRUZLOOQRWLQÀDWHDERYHPP+J
:KHQWKHPHDVXUHPHQWLVFRPSOHWHWKHDUPFXIIFRPSOHWHO\GHÀDWHV
Your blood pressure and pulse rate are displayed.
15
USING THE MEMORY FUNCTION
Every time you complete a measurement the monitor stores the blood pressure and pulse rate in the memory. The monitor automatically stores up to 14 sets of measurement values
(systolic pressure, diastolic pressure, and pulse rate). When 14 sets of measurement values are stored, the oldest record is deleted to save the most recent values.
1. Press the Memory (M) button.
The most recent set of measurement values appears on the display.
The values are displayed from the most recent to the oldest.
NOTE:
The number of the reading stored in the memory appears for a second before the pulse rate is displayed. The newest set is numbered “1”.
2. Press the START/STOP button to turn the unit off.
TO DELETE ALL THE VALUES STORED IN MEMORY
You cannot partially delete values stored in the memory. All values will be deleted.
1. Press the Memory (M) button.
Then while holding it down, press the START/STOP button down for more than 2 seconds until the display appears as illustrated below.
2. Press the START/STOP button to turn the unit off.
NOTE:
7KHPRQLWRUZLOODXWRPDWLFDOO\WXUQRIIDIWHU¿YHPLQXWHV
16
CARE AND MAINTENANCE
To keep your digital blood pressure monitor in the best condition and protect the unit from damage, follow the directions listed below:
Do not forcefully
bend the arm cuff or air tube. Do not fold tightly.
Clean the monitor with a soft dry cloth.
Do not use any abrasive or volatile cleaners.
Do not attempt to clean the cuff.
CAUTION
Do not submerge the device or any of the components in water. Do not subject the monitor to extreme hot or cold temperatures, humidity or direct sunlight.
CAUTION
Store the device and the components in a clean, safe location.
CAUTION
'RQRWVXEMHFWWKHPRQLWRUWRVWURQJVKRFNVVXFKDVGURSSLQJWKHXQLWRQWKHÀRRU
Remove the batteries
if the unit will not be used for three months or longer. Always replace all the batteries with new ones at the same time.
Use the unit consistent with the instruction provided in this manual.
Use only authorized parts and accessories.
Use the unit in a manner consistent with the instructions provided in this manual.
CAUTION
Use only Omron authorized parts and accessories. Parts and accessories not approved for use with the device may damage the unit.
CAUTION
&KDQJHVRUPRGL¿FDWLRQQRWDSSURYHGE\WKHPDQXIDFWXUHUZLOOYRLGWKHXVHU warranty. Do not disassemble or attempt to repair the unit or components.
17
SYMBOL
ERROR INDICATORS
CAUSE
Monitor could not detect pulse wave.
&XIIXQGHULQÀDWHG
Cuff not applied correctly.
&XIIRYHULQÀDWHG
CORRECTION
Take another measurement and remain still until the measurement is complete.
Remove the arm cuff.
Read “Taking a Measurement”.
Wait 2-3 minutes.
Take another measurement.
Batteries are worn.
Replace the four batteries. Refer to
“Battery Installation”.
Device error.
Contact Customer Service.
18
TROUBLESHOOTING TIPS
No power.
PROBLEM
No display appears on the unit.
CAUSES AND SOLUTIONS
Replace all four batteries with new ones.
Check the battery installation for proper placement of the battery polarities.
Measurement values appear too high or too low.
Blood pressure varies constantly. Many factors including stress, time of day, how you wrap the cuff, may affect your blood pressure.
Review the sections “Before Taking a Measurement” and “Taking a Measurement”.
19
FCC STATEMENT
Note:
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)
This product has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference in a residential installation. The product generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If the product does cause harmful interference to radio or television reception, which can be determined by turning the product on and off, the user is encouraged to try to correct the interference by one or more of the following measures:
5HRULHQWRUUHORFDWHWKHUHFHLYLQJDQWHQQD
,QFUHDVHWKHVHSDUDWLRQEHWZHHQWKHSURGXFWDQGWKHUHFHLYHU
&RQQHFWWKHSURGXFWLQWRDQRXWOHWRQDFLUFXLWGLIIHUHQWIURPWKDWWRZKLFKWKH receiver is connected.
&RQVXOWWKHGHDOHURUDQH[SHULHQFHGUDGLR79WHFKQLFLDQIRUKHOS
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)
This digital apparatus does not exceed the Class B limits for radio noise emissions from digital apparatus as set out in the interference-causing equipment standard entitled
“Digital Apparatus”, ICES-003 of the Canadian Department of Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables aux appareils numériques de Clase B prescrites dans la norme sur le materiel brouilleur:
“Appareils Numériques”, ICES-003 édictée par le minister des communications.
&KDQJHVRUPRGL¿FDWLRQVQRWH[SUHVVO\DSSURYHGE\WKHSDUW\UHVSRQVLEOHIRU compliance could void the user’s authority to operate the equipment.
20
LIMITED WARRANTIES
Your BP710 IntelliSense
®
Automatic Blood Pressure Monitor, excluding the arm cuff, is warranted to be free from defects in materials and workmanship appearing within
2 years from the date of purchase, when used in accordance with the instructions provided with the monitor. The arm cuff is warranted to be free from defects in materials and workmanship appearing within one year from the date of purchase when the monitor is used in accordance with the instructions provided with the monitor. The above warranties extend only to the original retail purchaser.
We will, at our option, repair or replace without charge any monitor or arm cuff covered by the above warranties. Repair or replacement is our only responsibility and your only remedy under the above warranties.
To obtain warranty service contact Customer Service by calling
1-800-634-4350
for the address of the repair location and the return shipping and handling fee.
Enclose the Proof of Purchase. Include a letter, with your name, address, phone
QXPEHUDQGGHVFULSWLRQRIWKHVSHFL¿FSUREOHP3DFNWKHSURGXFWFDUHIXOO\WRSUHYHQW damage in transit. Because of possible loss in transit, we recommend insuring the product with return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN
CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS
ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS
THAT MAY BE IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION
TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER
SPECIAL, INCIDENTAL, CONSEQUENTAIL OR INDIRECT COSTS,
EXPENSES OR DAMAGES.
7KLVZDUUDQW\SURYLGHV\RXZLWKVSHFL¿FOHJDOULJKWVDQG\RXPD\KDYHRWKHUULJKWV that vary by jurisdiction. Because of special local requirements, some of the above limitations and exclusions may not apply to you.
Visit our web site at:
Call toll free:
FOR CUSTOMER SERVICE
www.omronhealthcare.com
1-800-634-4350
21
Model:
Display:
Measurement Range:
Accuracy/Calibration:
Inflation:
Deflation:
Rapid Air Release:
Measurement Method:
Power Source:
Battery Life:
Operating Temperature /
Humidity:
Storage Temperature /
Humidity/Air Pressure:
Main Unit Weight:
SPECIFICATIONS
BP710 (HEM-7113-Z)
LCD Digital Display
Pressure: 0 to 299 mmHg,
Pulse: 40 to 180/min.
Pressure: ±3mmHg or 2% of reading
Pulse: ±5% of reading
Automatic by electric pump
Automatic pressure release valve
Automatic exhaust valve
Oscillometric method
1.5V 4 “AAA” batteries
Approx. 300 uses with 4 new alkaline batteries
50ºF to 104ºF (10ºC to 40ºC) /15 to 90% RH
Main Unit Dimensions:
Cuff Size:
Cuff Circumference:
-4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH
700-1060 hPa
Approximately 8 5/8 oz. (245g) not including batteries
Approx. 5"(l) × 4 1/8"(w) × 2 1/2" (h)
(128 mm × 104 mm × 64 mm)
Approximately 5 3/4"(w) × 17 1/2"(l) (146 mm × 446 mm)
Cuff tube 23 5/8" (600 mm)
Fits arm circumferences 9" to 13"
(220 mm to 320 mm)
22
SPECIFICATIONS
Memory:
Contents:
Up to 14 readings
Main Unit, Standard Adult Arm Cuff, Instruction
0DQXDODQG4XLFN6WDUW*XLGH
Large Cuff 13" - 17" (320 mm - 420 mm)
Optional Accessories:
= Type B
127(7KHVHVSHFL¿FDWLRQVDUHVXEMHFWWRFKDQJHZLWKRXWQRWLFH
23
Omron Healthcare, Inc.
1200 Lakeside Drive
Bannockburn, IL 60015 www.omronhealthcare.com
Copyright © 2010 Omron Healthcare, Inc.
Made in China 5325502-1B
Letöltés
Hirdetés
Gyakran ismételt kérdések
It is a digital blood pressure monitor that uses the oscillometric method to measure blood pressure.
The BP710 comes with the monitor, arm cuff, air tube, air plug, instruction manual, quick start guide, and standard adult arm cuff.
- Ensure you are relaxed and have not eaten, smoked, exercised, or consumed alcohol within the last 30 minutes.
- Sit in a comfortable chair with your back supported and feet flat on the floor.
- Place the arm cuff on your left or right upper arm, ensuring the cuff is level with your heart.
- Press the START/STOP button to begin the measurement.
- Remain still and silent during the measurement.
Displayed when the monitor detects an irregular rhythm two or more times during a measurement.
The Heartbeat Symbol appears with the measurement values if the reading is above 135 mmHg for systolic blood pressure or 85 mmHg for diastolic blood pressure.
According to the American Heart Association, the upper limit of normal for home blood pressure readings is 135 mmHg for systolic blood pressure and 85 mmHg for diastolic blood pressure.
No, the BP710 is not intended to be a diagnostic device. You should consult your physician for diagnosis and treatment.