Hewlett Packard 403B AC Voltmeter Operating and Service Manual
Mainos
Mainos
OPERATING AND SERVICE MANUAL
AC VOLTMETER
4038
, . S "tIS i
I f
I
J / un lUT
CI.VIlI,,'
FUNCTION
0"
ON
0
INPUT
HEWLETT~PACKARD
HEWLETT
=
iI
PACKARD
CE RTI FI CATION
The Hewlett-Packard Company certifies that this instrument was thoroughly tested and inspected and found to meet its published specifications when it was shipped from the factory. The Hewlett-
Packard Company further certifies that its calibration measurements are traceable to the U.S. National Bureau of Standards to the extent allowed by the Bureau's calibration facility.
WARRANTY AND ASSISTANCE
All Hewlett-Packard products are warranted against defects in materials and workmanship. This warranty applies for one year from .tne date of delivery, or, in the case of certain major components listed in the operating manual, for the specified period. We will repair or replace products which prove to
be
defective during the warranty period provided they are returned to Hewlett-
Packard. No other warranty is expressed or implied. We are not liable for consequential damages.
Service contracts or customer assistance agreements are available for Hewlett-Packard products that require maintenance and repair on-site.
For any assistance, contact your nearest Hewlett-Packard Sales and
Service Office. Addresses are provided at the back of this manual.
OPERATING AND SERVICE MANUAL
-hp- Part No. 00403-90010
MODEL 4038
AC VOLTMETER
Serials Prefixed: 0986A
If other serial prefixes require modification of this manual, the changes will appear in Appendix C or in an associated "Manual
Changes" supplement.
Copyright Hewlett-Packard Company 1962
P. O. Box 301, Loveland, Colorado, 80537 U. S.A.
Printed: January 1973
Mode1403B Table of Contents
I
Section
GENERAL INFORMATION . .
1-1. Description. : . . . .
1-5.
Instrument and Manual
1-7.
Accessories Available.
Section
II
Identification. • . . .
INSTALLATION . • . . . .
2-1.
Inspection . • . . • . .
2-4.
Power Requirements . . • • .
2-6.
Installation .
2-8.
Repackaging For Shipment.
TABLE OF CONTENTS
Section ill OPERATING INSTRUCTIONS
3-1.
Introduction . . . . . .
3-3.
Front Panel Description. .
3-5.
Operating Procedure. . . .
3-7.
Battery Charging Information.
3-9.
Instrument Temperature Limits. .
3-11. Input Protection . • . .
3-13. Voltage Measurements
3-18. Waveform Errors
3-21. Decibel Measurements
3-24. Impedance Correction Graph
3-27. Current Measurements . . . .
3-28.
Clip -On Probe. . . . . .
.
.
.
Page
1-1
1-1
1-1
1-1
Page
2-1
2-1
2-1
2-1
2-1
Page
3-1
3-1
3-1
3-1
3-1
3-1
3-1
3-1
3-1
3-4
3-4
3-4
3-4
Section
V MAINTENANCE
5-1.
Introduction
Section
VI REPLACEABLE PARTS . .
6-1.
Introduction.......
6-4.
Ordering Information. .
6-6.
Non-Listed Parts. . .
.
.
5-6.
Test Instruments Required.
5-8.
Meter, Mechanical Zero.
5-10. Performance Checks
5-12.
Calibration Checks .
.
5-13.
5-16.
5-17.
Frequency Response Checks.
Noise Check .
Input Resistance Check. .
5-18. Adjustment and Calibration
Procedures. . . . . . . . .
5-20.
5-21.
5-23.
Power Supply Adjustment
Input Resistance . . . . .
Overload Check. . .
5-24.
5-25.
Calibration and Tracking.
High Frequency Response
5-26.
30-Volt Response Calibration .
5-27. Troubleshooting . . . . . . . . . .
5-29. Repair. . • . . . . . . . . . . . . .
5-30.
Cabinet Removal .
5-31.
5-33.
5-35.
5-37.
Servicing Etched Circuit Boards
Transistor Replacement
Function Switch Repair
.
.
Fluorescent Indicator Decal.
Page
5-1
5-1
5-1
5-1
5-2
5-2
5-2
5-3
5-3
5-6
5-6
5-6
5-7
5-7
5-7
5-3
5-3
5-4
5-4
5-4
5-5
5-5
5-5 -
Page
6-1
6-1
6-1
6-1
Section
IV THEORY OF OPERATION . .
4-1.
Introduction . • . . . .
4-3.
Preliminary Attenuator .
4-6.
Input Circuit. • • . . .
4-11. Intermediate Attenuator .•
4-20. Meter Rectifier Circuit
4-25. Power Supply. . . • . . • .
Page
4-1
4-1
4-1
4-1
4-2
4-2
4-3
Appendix
A
B
C
CODE LIST OF MANUFACTURERS
SALES AND SERVICE OFFICES
MANUAL BACKDATING CHANGES
LIST OF
Number
1-1.
1-2.
3-1.
Specifications. . . . . . • . . . . . .
Accessories Available . • . . . • • . .
Effect of Harmonics on Model 403B
Page
. 1-1
.• 1-2
3-4
Voltage Measurements • . • • • • .
3-2.
Examples of Voltage and DB Measurements. 3-4
5-1.
Test Instruments Required. . . . . • . . • 5-1
TABLES
Number
5-2.
Calibration Check Table. . . . . . .
5-3.
Troubleshooting. . . . . . . . . .
5-4.
Test Procedure Troublesliooting .
5-5.
Transistor Replacement.
6-1.
Replaceable Parts.
6-2.
Mechanical Parts
Page
5-2
5-6
5-6
5-7
6-2
6-5
Number
1-1.
Model 403B Transistorized AC
LIST OF ILLUSTRATIONS
Page
Voltmeter. • . . . . . . .
. . . . . . 1-0
3-1.
Front Panel Description . . . . • . . . . . 3-2
3-2.
Voltage Measurements. . . . • . . . . . . 3-3
3-3.
Model403Blmpedance Correction Graph . . 3-5
4-1.
Model403BFunctionaiBIockDiagram . • . 4-0
4-2.
Input Amplifier. . . • . • . . . . . . .
. 4-1
4-3.
Diode Current Vs Diode Voltage 4-2
4-4.
Fixed Amplifier Block Diagram. . .
. 4-2
4-5.
Meter Rectifier, Simplified Diagram . . . 4-3
Number
4-6.
5-1.
5-2.
Meter Rectifier Circuit.
Model 403B Top View.
Model 403B Bottom View
5-3.
Overload Check Setup . . . • • • • • .
5-4.
Performance Check Setup. .
5-5.
5-6.
Frequency Response Setup
Range SWitch Detail . . .
Page
4-3
5-0
5-2
5-4
5-4
5-5
. • • •• 5-8
5-7.
5-8.
6-1.
Function Switch Detail. •
Schematic Diagram
Exploded View .
5-9/5-10
5-9/5-10
. . . . .
6-7 iii
Section I
1-0
'>'
\
\.\
\
\
,\\;)
-s
I
OECI8HS
'''1100
8
J
1 r ~ I
,
I
./
,
:x
~/
" H I
RANGE
VOLn oe
Model 403B
Ac Voltmeter standard
Model 403B
Ac Voltmeter
Option 01
Figure 1-1.
Model 403B
Ac Voltmeter
Model403B
Model403B Section I
SECTION I
GENERAL INFORMATION
1-1.
DESCRIPTION.
1-2.
The -hp- Model 403B Transistorized AC Voltmeter is a general purpose instrument that measures rms values of ac voltages in the 5 Hz to 2 MHz range.
The instrument has full-scale ranges from 1 mV to
300 volts (-72 dBmto+52dBm) in a 1,3,10 sequence.
The Model 403B meter face is calibrated with the upper scale in volts (rms); the Model 403B Option 01 meter face is calibrated with the upper scale in dB. Models
403B and 403B Option 01 are shown in Figure 1-1, and specifications are given in Table 1-1.
on the title page of this manual, change sheets supplied with the manual will define the differences between your instrument and the Model 403B described in this manual.
Some serial numbers may have a letter separating the two sections of the number. This letter indicates the country in which the instrument was manufactured.
1-7.
1-8.
ACCESSORIES AVAILABLE.
To increase the usefulness of your instrument, the following accessories are available:
1-3.
The Model403B operates from Nickel Cadmium batteries.
The instrument also includes a self-contained battery charger which operates on 115 or 230 volts ac.
a.
-hp- Model 11005A Line Bridging Transformer.
b.
-hp- Model 11039A Capacitive Voltage
Divider.
1-5.
INSTRUMENT AND MANUAL
IDENTIFICATION.
1-6. Hewlett-Packard uses a two-section serial number.
If the first section (serial prefix) of the serial number on your instrument does not agree with those
1-9.
c.
Table 1-2 provides information and use of the accessories mentioned above as well as other useful accessories.
Table 1-1.
Specifications
-hp- Model 10111A BNC-To-Binding Post
Adapter.
RANGE:
O.
001 to 300 volts rms full scale (12 ranges) in a
1, 3, 10 sequence.
FREQUENCY RANGE:
5 Hz to 2 MHz
ACCURACY:
% of Full Scale.
MODEL 403B ACCURACY SPECIFICATION
Frequency
10 Hz to 1 MHz
5 to 10 Hz and
1 to 2 MHz
o
to 50 0 C
o
to -20 oC
±2%
±5%
±8%
±8%
NOMINAL INPUT IMPEDANCE:
2 megohms; shunted by < 60 pF on O. 001-volt to O. 03-volt ranges, < 30 pF on O. 1-volt to
300-volt ranges.
OVERLOAD PROTECTION:
Fuse protected:
DC ISOLATION:
Signal ground may be ±500 Vdc from external chassis.
POWER SUPPLY:
4 rechargeable batteries (furnished).
40-hour operation per recharge (20 hours at -20 0 C), up to 500 recharging cycles (expected battery life of
20,000 hours).
Recharging circuit is self-contained and functions automatically when instrument is operated from ac line (115 or 230V ±10%
48 to 440 Hz, < 3 watts).
MODEL 403B OPTION 01 ACCURACY SPECIFI-
CATION
Frequency o to 50 0 C o to -20 o c
10 Hz to 1 MHz ±0.2 dB ±0.7 dB
5 to 10 Hz and
1 to 2 MHz
±0.4 dB ±0.7 dB
METER:
Individually calibrated, taut band.
Responds to average value of input waveform and is calibrated in the rms value of a sine wave.
TEMPERATURE RANGE:
-20oC to +50oC.
DIMENSIONS:
5-1/8 in. wide, 6-3/32 in. high, 8 in. deep.
ACCESSORIES AVAILABLE:
-hp- 11002A Test Leads, 5 ft. long, dual banana plug to alligator clips.
-hp- 11003A Test Leads,
5 ft. long, dual banana plug to probe and alligator clip.
ACCESSORY FURNISHED:
Detachable power cord.
1-1
Section I Model403B
Model No.
11002A
11003A
1l01lA
1l005A
1l039A
456A
11004A
Test Leads
Adapter
Use
Table 1-2.
Accessories Available
Features
Dual Banana Plug to Alligator Clips
Dual Banana Plug to Probe and Alligator Clip
Line Bridging Transformer
Provides balanced 600 -ohm input to unbalanced 600-ohm output for measurements on balanced lines.
Capacitive Voltage Divider
(Division ratio: 1000:1)
Binding Post to BNC
Terminating Resistance: 600 or 10 K ohms
Frequency Range: 20 Hz to 45 kHz
Power Handling Capacity: +15 dBm (4.5 V into 600n)
AC Current Probe
1 mY/rnA ±1% at 1 kHz
Line Matching Transformer
Provides balanced 135 or 600 ohm input to 600 ohm unbalanced output for measurements on balanced lines.
Accuracy:
Input Capacity:
Max. Voltage Rating:
±3%
15 pF ± 1 pF
60 Hz 25 kV, 100 kHz
22 kV, 1 MHz 20 kV, 10 MHz 15 kV, 20 MHz 7 kV
Resistance Max. Current Accuracy
< 50 milliohms 1 amp rms
1. 5 amp peak
±2%
100 Hz to 3 MHz
Terminating Resistance: 600 or 10 K ohms
Frequency Range: 5 kHz to 600 kHz
Power Handling Capacity: 22 dBm (10 V into 600 n)
1-2
Model403B section II
2-1.
INSPECTION.
2-2.
Unpack the instrument upon receipt and inspect it for signs of physical damage such as scratched panel knobs etc.
If there is any apparent damage, file a claim with the carrier and refer to the warranty page on the back side of the front cover of this manual.
2-3.
An electrical inspection should be performed as soon as possible after receipt.
To aid in electrical inspection, performance checks are included in section V, Paragraph 5-10.
2-4.
POWER REQUIREMENTS.
2-5.
The ac power circuit which provides charging current tothe Nickel-Cadmium batteries in the instrument can be connected to a 115- or 230- volt, 48 - 440·
Hz, source. A switch located on the rear panel of the instrument allows the user to select 115- or 230volt modes of operation.'
I
WARNING
I
WHEN THE 403B IS USED WITHOUT
THE POWER CORD, THERE IS NO
GROUND RETURN.
USE CAUTION
TO AVOID ELECTRICAL SHOCK.
SECTION II
INSTALLATION
2-6.
INSTALLATION.
2-7.
The -hp- Model 403B is fully transistorized; therefore no special cooling is required.
However, the instrument should not be operated where the ambient temperature exceeds 50 0 C (131 0 F).
2-8.
REPACKAGING FOR SHIPMENT.
- - - - - - N O T E - - - - -
If the instrument is to be shipped to
Hewlett-Packard for service or repair, attach a tag to the instrument identifying the owner and indicating the service or repair to be accomplished; include the model number and full serial number ofthe instrument.
In any correspondence, identify the instrument by model number, serial number and serial number prefix.
2 -9.
The following is a general guide for repacking for shipment.
If you have any questions, contact your local -hp- Sales and service Office.
(see Appendix B for office locations.) a.
Place instrument in original container if available.
If original container is not available, one can be purchased from your nearest -hp-
Sales and service Office.
If original container is not used, b.
Wrap instrument in heavy paper or plastic before placing in an inner container.
c.
Use plenty of packing material around all sides of instrument and protect panel faces with cardboard strips.
d.
Place instrument and inner container in a heavy carton or wooden box and seal with strong tape or metal bands.
e.
Mark shipping container with 'Delicate Instrument, " "Fra~ile" etc.
2-1
Model403B section ill
SECTION III
OPERATING INSTRUCTIONS
3-1.
INTRODUCTION.
3-2.
This voltmeter is ready for use upon receipt from the factory and will give specified performance after a short warm-up period.
Allow approximately
60 seconds warm-up for optimum performance.
3-3.
FRONT PANEL DESCRIPTION.
3-4.
A description of front panel controls is given in
Figure 3-1.
The descriptions are keyed to the photo that accompanies the figure.
3-5.
OPERATING PROCEDURE.
3-6.
The operating procedure for the Model 403B is given in Figure 3-2.
Instructions are keyed to the photo that accompanies the figure.
3-7.
BATTERY CHARGING INFORMATION.
3-8.
The 403B has a self-contained battery charger.
This instrument is continually charging the batteries whenever the FUNCTION switch is ON and the line cord connected to a 115 or 230 volt source.
In the event of complete discharge, the 403B can be used after twenty minutes of recharging with the line cord connected to an ac source. Complete recharge requires approximately 60 hours (depending on setting of R39) when the Nickel Cadmium cells are completely discharged.
(Refer to Section IV, Paragraph 4-25).
THE FOUR NICKEL CADMIUM BAT-
TERIES IN THE -HP- 403B ARE IN
HERMETICALLY-SEALED CONTAIN-
ERS. UNDER HIGH TEMPERATURES
(ABOVE 50 0 C), HYDROGEN IN THE
HERMETICALLY -SEALED BATTERY
CONTAINER CAN BUIlD UP TO LARGE
PRESSURE, CAUSING DAMAGE TO THE
BATTERIES AND/OR INSTRUMENT.
(REFER TO SECTION IV, PAGE 4-3.)
3-9.
INSTRUMENT TEMPERATURE LIMITS.
3-10.
This instrument has certain temperature limitations.
The design of this instrument has provided for safe and stable operation over the range of -20 to
+50 o C (-4 to +131 0 F).
This temperature range is quite adequate for most users; however, keep these limits in mind when operating under field conditions.
Internal temperatures in excess of 122 0 F are quite easy to obtain if the instrument is left in the sun, even if the air temperature is quite moderate.
A good practice is to be certain that the instrument is not stored or operated in direct sunlight to avoid the possibility of reduced performance.
When using 403B at temperatures below OOC, be certain the batteries are fully charged prior to subjecting instrument to this temperature.
3-11.
INPUT PROTECTION.
3-12.
A 1/16 A fuse is included in series with the input Circuit which will open with repeated or excessive overload.
This fuse is accessible when the cabinet is removed.
A spare fuse is included inside the instrument.
3-13.
VOLTAGE MEASUREMENTS.
3-14.
Always leave the instrument onthe1-voltrange or higher when making initial connections to circuits which have dc levels over 25 volts.
Then switch to the appropriate lower range to obtain an up-scale reading.
This practice should be used when making power supply ripple measurements where the dc voltage may be as much as 600 volts, but the ac ripple is only a few millivolts.
To obtain specified accuracy, the ground strap should be connected except when the signal ground is dc offset from chassis ground.
3-15.
If measurements are made from a high-impedance source, hum pick-up can affect the meter indication because of high impedance of both the source and voltmeter.
Shielded leads will reduce pick-up although they will cause an increase in the capacity shunted across the source, with the possibility of excessive circuit loading.
3-16.
The rated 2 megohms input resistance will be effectively reduced (above 1 kHz) by shunt input capacity. (This fact is true for any ac voltmeter.) 50 pF has a reactance of 0.8 megohm at 4 kHz, 80,000 ohms at 40 kHz, etc. The shunt capacity decreases on the higher ranges (see Table 1-1). This factor should be considered when measuring higher frequency voltages in circuits of moderate impedance level.
3-17 .
severe RF circulating currents are generated at potentials approaching 300 volts in the 1 to 2 MHz frequency range. These severe ground currents limit the accuracy of the 403Bto±10% on the 300-volt range.
By using -hp- accessories 1000lA (10:1 divider) and a 10111A (adapter) shunted by a 2-megohm resistor, the accuracy of the 403B can be retained to ±5%.
3-18.
WAVEFORM ERRORS.
3-19.
In order to maintain accuracy of measurement, one must remember that this instrument is an average responding device, but the meter scale is calibrated in terms of the rms value of a pure sine wave.
If the waveform of the voltage being measured contains appreciable harmonics or other spurious voltages, the meter indication will deviate from the true rms value on the order indicated by Table 3-1.
3-1
section
m
Mode1403B
.Ull
lilt
3-2
CD
FUNCTION: Three-position switch checks battery charge when the switch is in the ON position; it applies 27. 5 volts from the Nickel-
Cadmium batteries to the Mode1403B circuitry.
When in the BATT. TESf position, the meter should read above 2.4 volts on the "0-3" meter scale, which is equivalent to 24 Vdc at the battery.
CD
BATT. CHARGE: Glows when instrument is connected to an AC source with the FUNCTION switch turned to the ON position.
Fluorescent Iooicator: Glows when instrument is on.
RANGE: selects range from 0.001 volt to 300 volts rms full scale in a 1, 3, 10 sequence.
CD
INPUT: Connect voltage to be measured to these terminals.
Figure 3-1.
Front Panel Description
Model403B Section m
4058-A-IIIZA o
Select range which gives a reading in the upper
2/3 of the meter scale (This will insure the highest degree of accuracy).
CD
Connect voltage to be measured at the INPUT terminals (red terminal positive and ungrounded black terminal negative).
- - - - N O T E - - - - -
The outer black ground terminal is connected to chassis ground.
For voltage measurements at chassis ground, connect the ungrounded black terminal to the grounded black terminal.
Set FUNCTION switch to BATT. TEsr; front panel meter reading should be greater than
2.4 volts (corresponds to 24 volts at battery).
If less than 2.4 volts, the battery needs recharging.
Refer to Paragraph 3-7 for battery charging instructions.
Set FUNCTION switch to ON; lamp function knob will glow, and the instrument is ready for voltage measurement.
When the instrument is connected to an AC power source and the FUNCTION switch is to
ON, the BATT. CHARGE lamp will glow, indicating that the battery is being charged.
Figure 3-2.
Voltage Measurements
3-3
Section ill Model403B
Table 3-1.
Effect of Harmonics on Model 403B
Voltage Measurements
Input Voltage
Characteristics
Fundamental = 100
Fundamental +10%
2nd harmonic
Fundamental +20%
2nd harmonic
Fundamental +50%
2nd harmonic
Fundamental +10%
3rd harmonic
Fundamental +20%
3rd harmonic
Fundamental +50%
3rd harmonic
True RMS
Value
100
100.5
102
112
100.5
102
112
Value Indicated by 403B
100
100
100 - 102
100 - 110
96 - 104
94 - 108
90 - 116
3-20.
This table is a general one and applies to any average responding rms calibrated voltmeter.
As can be seen in the table, errors are small even with a badly distorted signal (e. g.; 20% 2nd harmonic gives
+0, -2% error).
3-21.
DECIBEL MEASUREMENTS.
3-22.
Measurements in terms of decibels are made in the same way as voltage measurements except that the indication is read on the dB scale (-12 to +2 dB).
The decibel level is the algebraic sum of the meter dB scale indication and DB VOLTS (RANGE) position.
3-23.
To read power directly in dBm, (0 dBm=l milliwatt into 600 ohms) the mE;asurement must be made across 600 ohms.
Comparative dB measurements
(without respect to reference level) may be obtained by direct reading provided each measurement is made across the same impedance value.
The difference in decibels between two or more measurements may be obtained by reading directly from the dB-scale indications.
(For examples of dB measurements, refer to Table 3-2.)
3-24.
IMPEDANCE CORRECTION GRAPH.
3-25.
To obtain the level in dBm with respect to impedances other than 600 ohms, the meter correction.
graph shown in Figure 3-3 may be'Used.
The level
In dBm will be the algebraic sum of the level as indicated on the meter and the correction shown on the graph.
For example, if the range switch is at the
+30 dB position, the measurement made across 90 ohms, and the indication ontheDB scale +1, the level in dBm is obtained as follows:
+ 1 (dB-scale indication)
+30 (range switch position)
+31 (level in dB as indicated by meter)
+ 8 (correction for 90-ohm impedance)
+39 dBm
3-4
Table 3-2.
Examples of Voltage and DB
Measurements
Range
Switch
Meter
Scale
Meter
Indicates
Actual
Level
Voltage measurements:
300
10
.003
.001
3
1
3
1
DB measurements:
+40 dB
+40 dB
+10 dB
-30 dB
-30 dB
*-50 dB
-60 dB dB dB dB dB dB dB dB
1.8
0.44
2.3
.27
+2 dB
-7 dB
-6 dB o dB
-8 dB
-9 dB
+1 dB
180
4.4
.0023
.00027
+42 dB
+33 dB
+ 4 dB
-30 dB
-38 dB
-59 dB
-59 dB
*NOTE: In cases where a meter scale reading below -8 dB is obtained, it is best to switch to the next lower range on the instrument so a reading will be obtained in the upper portion of the scale where highest accuracy may be obtained.
The same situation exists for voltage measurements.
When a reading is obtained in the lower 1/3 scale, the range switch should be switched to the next lower range to obtain a reading in the upper 2/3 scale.
3-26.
ment made across 10,000 ohms:
3-27.
CURRENT MEASUREMENTS.
3-28.
For the same conditions, with the measure-
+ 1 (dB-scale indication)
+30 (range switch position)
+31 (level in dB as indicated by meter)
-12.5 (correction for 10,OOO-ohm impedance)
+18.5dBm
CLIP-ON PROBE.
3-29.
The -hp- Model 456A Current Probe provides quick measurement of current from 1 rnA to 1 amp full scale with minimum circuit loading.
3-33.
To use the Model 456A, simply clamp the probe around the current carrying wire and plug the output into the Model 403B.
The probe output is 1 mV/mA.
Model403B
Section III
+20
+15
Z
0
~
U
W a: a:
0 u a: w
~
W
~
~
OJ
0
I
+10
+5
0
-5
-10
-15
-20
-25
I 10 100 IK 10K
IMPEDANCE - OHMS lOOK
Figure 3-3.
Model 403B Impedance Correction Graph
3-5
Model403B
Section IV
SECTION IV
THEORY OF OPERATION
4-1.
INTRODUCTION.
4-2.
The Model403B includes a preliminary input attenuator, a high impedance emitter follower circuit, a range attenuator and a wide range fixed gain amplifier.
Refer to Figure 4-1.
4-3.
PRELIMINARY ATTENUATOR.
4-4.
The RANGE switch is divided up into two sections: the preliminary attenuator, located between the input terminals and Q1, and the intermediate attenuator, located between Q2 and Q3. The preliminary input attenuator has two ranges, 100:1 and 10,000:1, which are switched in at the appropriate time to keep the input voltage to Q1 less than O.
030 volt.
This not only prevents overloading the input system, but also provides the necessary accurate attenuation to work with the intermediate attenuator to produce the conventional 1, 3, 10 sequence for correct meter operation.
4-5.
The attenuators are of the compensated resistorcapacitor (rc) type, with the capacitive division ratio made equal to the resistive ratio to maintain a constant division ratio at all frequencies.
By making one of the capacitors adjustable, the small variations in stray circuit capacity can be compensated for, so the voltmeter will have a flat response.
The exact division ratio is set at low audio frequencies by the trimmer potentiometers, which bring the resistor division ratio to the exact value.
-13V
R7
C9
+
4-6.
INPUT CIRCUIT.
4-7.
Rll, CR1, and CR2 make up a limiting circuit which is used for overload protection to prevent high instantaneous voltages from being impressed on the base of Q1.
F1 is a 1/16 amp fuse used to protect the 403B against a continuous or repeated overload.
4-8.
Since transistors are inherently low impedance devices, a need for a high input impedance is required.
Referring to Figure 4-2, it would seem that the input resistance of the first stage would be approximately
Ri of a grounded collector configuration in parallel with R9, plus the R7-RB combination.
Q1 and Q2 ~e emitter followers, exhibiting unity gain and no phase reversal.
(Ri
=
approx. input Z of a common collector stage.)
4-9.
The output of Q2 is fed back to the junction of
R9 and R7 -R8. There is an ac voltage existing at this point that is very nearly the same amplitude as, and in phase with, the input voltage. Since a very small ac voltage exists across R9 (due to the feedback from
Q2), the input current lin will be very small. Thereby:
Zin
E in
=
~
10
It can be seen that when lin is very small, the apparent Zin becomes very large.
4-10.
The Ri of Q1 is increased in a similar manner by feeding the Q2 emitter voltage to both the collector and emitter of Q1.
-13V
CIO
RI5
02
-S-POSITIVE
FEEDBACK
+ ~ ~
RIG
~EOUT
CI2
+
Figure 4-2.
Input Amplifier
4-1
4-2
Section IV
Mode1403B
4-11.
INTERMEDIATE ATTENUATOR.
4-12.
The output of Q2 is fed to the intermediate section of the range attenuator.
The range attenuator is a voltage divider, in sequence with the preliminary attenuator.
A (1, 3, 10 etc.) ratio is obtained resulting in 10 DB steps.
Refer to Figure 4-1.
4-13.
Refer to schematic diagram (Figure 5-8) in the back of this manual.
4-14.
Transistors Q3 through Q6 make up the fixed gain amplifier which is used to develop the current for (full scale) meter deflection and to provide the meter circuit with a high impedance source for linear operation at all current levels.
4-15.
The output of the intermediate range attenuator is fed to the base of Q3 (djfferential amplifier), and compared with a feedback signal to its emitter from the meter circuit.
This difference signal is fed to Q4 which in turn is directly coupled to Q5 and Q6.
Q4 is a grounded emitter amplifier.
Q5 is a common collector amplifier which impedance matches Q6, a common base amplifier.
The direct couple feature of the amplifiers is necessary because of the low-frequency
(5 hertz) response of the 403B.
R24 through R26 make up the dc feedback loop which tends to minimize any tendency for dc drift due to ambient temperature change.
R33 corrects the total gain of Q3 through Q6.
4-16.
The meter source impedance is increased by the use of negative feedback from the output of the meter rectifier bridge to the emitter of Q3. Resistor
R28 through R30, and C15 and C16 correct the phase of the feedback at high frequ~ncies.
4-17.
The necessity of high meter source impedance can be understood by referring to Figures 4-3 and 4-4.
4-18.
To have correct voltmeter action, it is necessary that the change in meter circuit be proportional to a change in amplifier input voltage.
The load resistance, then, should remain constant. Note in Figure 4-3, however, that when 1
0
(and therefore the diode voltage Ed) decreases, the diode resistance Rei
(and therefore the load resistance) increases, affecting meter linearity.
Note in Figure 4-4 that
Rei appears in series with Ro' the source impedance. The effect on output current, due to changes in diode resistance with voltage, can be minimized by feeding the meter circuit from a constant current or high impedance source.
In this way, changes of diode resistance will have a negligible effect ,on the total current passing through them and hence through the meter.
4-19.
The effect of diode resistance change is further minimized by Q6 current through R35 which impresses a fixed 0.3 volt across CR3 and CR4, biasing them close to their contact potential.
~I o
/
/
V
/
V
/
0.6
/
VOLTAGE INCREASING----•• RO
DIODE VOLTAGE (Ed)
Figure 4-3.
Diode Current Vs Diode Voltage
_ _ _ NEGATIVE D.C.
FEEDBACK RZ4
' - - - - - - - - - - - - - . _ _ - - - i M
+---NEGATIVE A.t. FEEDBACK
T C20
Figure 4-4.
Fixed Amplifier Block Diagram
4-20.
METER RECTIFIER CIRCUIT.
4-21.
The meter rectifier circuit is arranged in a bridge-type configuration, with a crystal diode and a capacitor in each branch and adc microammeter connected across its midpoints. The current through the meter is proportional to the average value of the input voltage waveform.
4-22.
The 403B meter rectifier circuit operation can best be explained by analyzing the circuit in a simplified
Model403B
Section IV
CRI
+
I
I
I
•
1
1 t-----------,
I
I
•
+
CR2
R35
---,
I
I
I
I
I
I
I
I
L ___
R o f\.;
E s
___ J
I
I
1
1
I
1
1
I
1
1
CI7
+
CR3
R m
CI9
+
CR4
+
R36
+13
+
CIS
E f
TO 0 3 - - -....
'~~OA-8-0'a9
Figure 4-5.
Meter Rectifier, Simplified Diagram form.
Figure 4-5 shows a voltage source generating a voltage Es across a circuit made up of CR1, CR2,
M1, Rf, and C1, C2.
Note that the current flow for each half cycle (as imicated by the arrows) always passes through the meter in the same direction.
4-23.
In this circuit, disregarding contact potential aM assuming zero meter resistance, the circuit could be considered as a small resistance made up of CR1 aM CR2, in series with one capacitor (C1 + C2) in series with across
Rt.
Therefore, there will be a voltage
Rf
proportional to the input signal.
4-24.
In the actual 403B meter rectifier circuit, capacitors C17 and C18 provide a path for the AC feedback loop.
The generator (Q3-Q6) with its large internal impedance (Ro) develops a voltage across the bridge.
The meter is deflected according to the average value of the input: voltage.
The signal across
Rt
as in Figure 4-6 provides negative feedback, resulting in extreme linear meter operation and large
Ro.
4-25.
POWER SUPPLY.
4-26.
The Model 403B operates on batteries only.
This instrument uses four 6. 5 volt nickel cadmium batteries and is designed to have a battery life of 40 hours before recharging.
4-27.
R39 has been adjusted at the factory for a charging rate of 6.2 rnA to prolong battery life.
If the instrument is used frequently in the field, R39 can be adjusted for a charging rate of 11 rnA.
IF R39 IS ADJUSTED TO THE 11 rnA RATE, THE
INSTRUMENT SHOULD BE USED ON BATTERIES
ONLY EXCEPT WHEN RECHARGING BATTERIES.
THE BATTERIES ARE RECHARGED WHEN THE
403B IS CONNECTED TO AN AC SOURCE WITH
THE FUNCTION SWITCH TO ON. THE BATTERY
LIFE OF THE INSTRUMENT CAN BE PROLONG-
ED AT THE 11 rnA CHARGING RATE IF THE
INSTRUMENT
CHARGED.
lS NOT CONTINUOUSLY OVER-
Figure 4-6.
Meter Rectifier Circuit
4-28.
When the function switch is in the BATT TEST position, and the instrument indicates a battery voltage of 2.4 volts, recharge the batteries for 20 to 25 hours at the 6.2 rnA rate to completely recharge the batteries in the instrument.
A longer charging period will be required if the batteries have been allowed to discharge below 24 volts.
4-29.
Figure 5-8 illustrates the Power Supply
Assembly, which supplies 4.8 rnA of current to the instrument and 6.2 rnA of current to the batteries. R39 is used to control the amount of current used to charge the batteries and caution must be used if R39 is adjusted to maximum charging rate.
THE FOUR NICKEL-CADMIUM BATTERIES
IN THE 403B ARE IN HERMETICALLY
SEALED CONTAINERS.
THE USER MUST
BE COGNIZANT OF TEMPERATURE EX-
TREMES WHILE CHARGING THE BATT-
ERIES.
UNDER HIGH TEMPERATURE
(ABOVE 50
0
CENTIGRADE), HYDROGEN IN
THE HERMETICALLY SEALED BATTERY
CONTAINER CAN BUILD UP LARGE PRES-
SURE CAUSING DAMAGE TO THE BATTER-
IES AND/OR INSTRUMENT.
DO NOT CHARGE
BATTERIES ABOVE 40
0
CENTIGRADE OR
104
0
FAHRENHEIT, IF R39 IS SET ABOVE
6. 2 MA CHARGING RATE.
4-30.
Figure 5-8 illustrates a conventional power supply.
For 115 volt operation, the power transformer primaries are connected in parallel, and in series when used for 230 volt operation.
The rectifier circuit is a conventional full wave bridge using C21 for a filter capacitor.
Diode CR9 (7 volt breakdown diode) am Q7 make up the Constant Current Generator.
The collector current of Q7 is equal to the voltage across CR9 divided by R37 and R39.
4-31.
CR10 prevents the batteries from discharging to the charging circuit when the instrument is in the
OFF position.
4-3
Section V Model403B
METER METER
(-) (+) -13V -6.5V
t
l3V
CI6
·CAL IMV
2MC
R29
METER CAL
IMV RANGE
403B-A-1333
PRE-AMPL SIG PRE-AMPL
INPUT GRD OUTPUT
R33
LOOP GAIN
ADJ
METER AMPL
INPUT
Figure 5-1.
Model 403B Top View
5-0
Model403B section V
SECTION V
MAINTENANCE
5-1.
INTRODUCTION.
5-2.
This section contains test and maintenance information for your 403B.
Included is a quick performance check that may be made with the instrument in its cabinet, as a part of roUtine maintenance or as a part of incoming quality control inspection.
5-3.
This instrument should require very little maintenance.
Should failure occur, however, a troubleshooting paragraph (5-27) has been included to assist in locating the failure.
An exploded view of the Model
403B is given in Figure 6-1 to help in locating parts.
5-8.
METER, MECHANICAL ZERO.
5-9.
When the meter is properly zero-set, the pointer will rest over the zero mark on the meter scale when the voltmeter is 1) at normal operating temperature, 2) in its normal operating position, and
3) turned off.
Zero-set as follows to obtain best accuracy and mechanical stability.
a.
Allow the voltmeter to operate for at least
20 minutes; this allows the meter movement to reach normal operating temperature.
b.
Turn voltmeter off and allow 30 seconds for all capacitors to discharge.
c.
Rotate mechanical zero-adjustment screw
CLOCKWISE.
5-4.
If it becomes apparent that transistor replacement is necessary, consult Paragraph 3-33 to avoid damage to the new transistor. Care should be taken not to damage the printed circuit board.
5-5.
Errors may be introduced in the 403B because of the capacity added in the circuit after cabinet replacement. Therefore, after making gain or frequency response adjustments) temporarily place covers back on instrument and recheck the adjustment.
d.
Continue to rotate adjustment screw CLOCK-
WISE; STOP when pointer is right on zero.
e.
Rotate adjustment screw COUNTERCLOCK-
WISE 15 0 to remove tension.
5-6.
5-7.
TEST INSTRUMENTS REQUIRED.
Table 5-1 gives the test equipment required to check the 403B.
f.
If step e causes the meter to move off zero, repeat procedure, making the
COUNTERCLOCKWISE rotation less than
15 0 •
Table 5-1.
Test Instruments Required
INSTRUMENT TYPE
DC Electronic Voltmeter
MINIMUM REQUIRED SPECIFICATIONS sensitivity: 1 volt full scale minimum
Input resistance: 10 megohms or higher
RECOMMENDED
-hp- INSTRUMENTS
Model 412A
Vacuum Tube Voltmeter
Voltmeter Calibration
Generator
Output voltage range: .001 to 300 volts
Signal frequency: 400 Hz
Distortion: less. than 0.2%; Accuracy
±0.25%
Model 738BR
Voltmeter Calibrator
Termination
Resistor
Resistor
AC Electronic Volt.meter
Clip OnDC
Milliammeter
Oscillator
Feedthrough, 50' ohm
15 kilohm ± 10%,' 10 watt
200 kilohm ± 1%
~
Input impedance: 10 megohms shunted by
25 pF (below the 0.3 volt range)
Accuracy: ±2% from 20 Hz to 1 MHz
Current Range: 3 mA to 1 ampere
Accuracy: ±3% ±0.1 mA
Model 11048B
0811-9034
0757-0128
Model 400E/EL Electronic
Voltmeter
Model 428A/B
DC Milliammeter
Model 652A Frequency range: 5 Hz - 2 MHz
Output impedance: 600 ohms, 50 ohms
Output level: o to 1 V rms into 600 ohms
Monitor: meter circuit with expand function
5-1
Section V Model403B
403B-A-1334
51
RANGE
R3 R4
5-10.
5-11.
The performance checks are in-cabinet tests that compare the 403B with its specifications.
These procedures can be used bothfor incoming and periodic inspections. Refer to Table 5-1 for test equipment required throughout this procedure.
5-12.
a.
b.
c.
d.
e.
PERFORMANCE range.
306 V.
CHECKS.
CALIBRATION CHECKS.
Figure 5-2.
Model 403B Bottom View
- - - - - N O T E - - - - - -
Do not adjust the REF SET control for the remainder of the check.
Rotate the 403B 'FUNCTION switch to BATT
TEST.
Meter' should read 2.4 volts on the
3.0 volt scale.
If 403B does not read 2.4
volts, recharge the batteries.
Connect the 403B as shown in Figure 5-4.
Set the Voltmeter Calibrator for 400 Hz and
300 V rms and set the 403B to the 300 V
The 403B should indicate between 294 V and
Repeat steps c and d for each voltage indicated in Table 5-2.
Table 5-2.
Calibration Check Table
VOLTMETER
CALIBRATOR MODEL 403B
OUTPUT (400 CPS RMS) INDICATION (V RMS)
Minimum Maximum
300V
100V
30V
10V
3V lV
0.3V
O.lV
30mV
10mV
3mV lmV
294V
98V
29.4V
9.8V
2.94V
0.98V
0.294V
0.098V
29.4mV
9.8mV
2.94 mV
0.98 mV
306V
102V
30.6V
10.2V
3.06V
1.
02V
0.306V
0.102V
30.6 mV
10.2 mV
3.06 mV
1. 02 mV
5-13.
FREQUENCY RESPONSE CHECKS.
5-14.
HIGH FREQUENCY RESPONSE CHECK.
a.
Connect the 403B as shown in Figure 5-5.
b.
Set the 652A for 400 Hz and adjust the output level for 0.94 mV indication on 1 mV range of the 403B meter.
c.
Place the 652A OUTPUT MONITOR switch to
EXPAND and adjust the REF SET control for a convenient reference point.
d.
Change the 652A frequency to 500 kHz and adjust the output level for the same reference level as in step c.
e.
The 403B meter indication should be within
0.92 mV and 0.96 mY.
f.
Repeat step d for 1 MHz and 2 MHz. The 403B meter reading should be between O. 89 mV and 0.99 mY.
g.
Repeat the same procedure for the 0.003,
0.01, 0.03, 0.1 and 3 volt range.
5-2
Mode1403B Section V h.
Turn the 403B RANGE switch to 30 volts.
i.
j.
Connect an oscillator (-hp- Model 200 CD}to the 403B INPUT. Set the frequency to 400Hz.
Connect on AC Voltmeter (-hp- Model 400E
E L) to the oscillator output.
k.
Adjust the oscillator for a 20. 0 volt reading on the 403B meter. Record the reading on the monitoring voltmeter (connected in Step j).
1.
Adjust the oscillator frequency to 300 kHz.
Adjust the oscillator AMPLITUDE control until the monitoring voltmeter indicates the reference level recorded in step k.
m.
The 403B meter should indicate between 19. 4V and 20. 6V.
5-15.
LOW FREQUENCY RESPONSE CHECK.
a.
Connect the 403B as shown in Figure 5-5.
b.
Set the 652A for 400 Hz and adjust output level for 0.94 mV indication on 1 mV range of 403B.
c.
Place OUTPUT MONITOR switch on 652A to
EXPAND and adjust REF SET control for a convenient reference point.
- - - - NOTE - - - - - - -
Do not adjust the REF SET control for the remainder of the check.
d.
Change the 652A frequency to 5 Hz and adjust the OUTPUT AMPLITUDE for the same reference level as in Step c.
e.
The 403B meter indication should be between
0.89 mV and 0.99 mY.
f.
Repeat the same procedure for the 0.003,
0.01, 0.03, 0.1, and 3 volt range.
5-16.
NOISE CHE CK.
a.
Disconnect the 403B from the setup shown in
Figure 5-4 and rotate the 403B FUNCTION switch to ON.
b.
Terminate the 403B input with a lOOK ohmshielded load.
The 403B meter deflection should be less than 3% with battery operation and less than 8% on any range with ac operation.
5-17.
INPUT RESISTANCE CHECK.
a.
Connect the 403B as shown in Figure 5-4
(Position B).
b.
Rotate the 403B RANGE switch to 0.01.
c.
Adjust Voltmeter Calibrator for an output of
0.01 volt 400 Hz rms.
d.
The 403B meter should indicate between
9.0 mV and 9.5 mY.
- - - - - N O T E - - - - - - -
This corresponds to an input resistance of 1.
8 to 3. 8 megohms where:
Rinput =
E meter
-----:;::...:...;~---
E input - E meter x R series
5-18.
ADJUSTMENT AND CALIBRATION
PROCEDURES.
5-19.
The following is a complete adjustment and calibration procedure and should be made only if it has been definitely determined that the 403B is out of adjustment. Transistor changes are usually not cause for complete adjustment (see Table 5-5).
If the instrument fails any of the limits given in the following steps, carefully recheck your connections and procedure.
If the instrument still fails the step, refer to Table 5-3 and 5-4 for possible cause and corrective action.
- - - - - - N O T E - - - - - -
In order to avoid the effects of hand capacity, a tuning wand with a nonmetallic shank should be used for all adjustments.
5-20.
POWER SUPPLY ADJUSTMENT.
a.
Connect the 403B to a variable line transformer.
Set the line voltage to 115 volts; turn on the AC Voltmeter, and allow five minutes for warmup.
b.
Connect a Clip-On DC Ammeter (-hp- Model
428A/B) probe around the violet wire connected to battery BT4.
Adjust R39 (see Figure 5-2) for an indication of 6: 2 mA on the
DC Ammeter.
- - - - - NOTE - - - - - -
If the instrument is to be used frequently in the field, R39 can be adjusted for a fast charging rate of
11 mAo Do not charge batteries at temperatures above 40 0 C if R39 is set for 11 mA charging rate. Batt~ry life will be prolonged at the lower charging rate.
c.
Vary the input line voltage from 103 to 127 volts; the Clip-On DC Ammeter reading should not vary more than 1.
0 mA from the reference setting in step b.
d.
Set line voltage to 115 volts. Using a NEMA three prong to two prong ac adapter for isolation, connect an AC Voltmeter (-hp- Model
400E/EL) across the red (BTl) and violet
BT4} wires connected to the batteries; the ripple voltage should not exceed 1.
5 mV.
e.
Set 403B FUNCTION switch to OFF; disconnect ac power source and set FUNCTION switch to ON.
5-3
Section V f.
Connect volts probe of a DC Voltmeter (-hp-
Model 412A) to red wire connected to battery
BT 1; connect common lead of dc voltmeter to violet wire connected to battery BT4.
Model403B o o
VOLTMETER CALIBRATOR hp738BR o
©
oo@
0 o hp403B
AC VOLTMETER
' - - - - - - - - - - - - - - E : r < : : : : : : l A
200K
:t1"4 B
DC VOLTMETER MUST BE ISO-
LATED FROM 403B GROUND.
g.
Rotate 403B FUNCTION switch to BATT
TEST.
If DC Voltmeter reading is not 24 volts, recharge batteries in the 403B.
(See
Paragraph 3-7). Adjust R45 for a 403B-meter indication equal to voltage indicated on DC
Voltmeter.
5-21.
INPUT RESISTANCE.
5-22.
Check the 403B input resistance by following the procedure outlined in Paragraph 5-17.
If the input resistance is not within test limits, the value of
R6 should be changed (typical range of R6 is from 3.9
to 10 megohms).
o o
VOLTMETER CALIBRATOR hp738f3R o
©
o hp403B
AC VOLTMETER
/
Figure 5-3.
Overload Check Setup
5-23.
OVERLOAD CHECK.
a.
Connect the 403B as shown in Figure 5-3.
THE 15KQ RESISTOR MUST BE
CONNE CTED AS SHOWN IN FIG-
URE 5-3 TO PREVENT DAMAGE
TO VOLTMETER CALIBRATOR.
b.
Rotate the 403B FUNCTION switch to OFF; connect anAC Voltmeter (-hp-ModeI400E/EL) between the base of Q1 and chassis ground.
c.
Rotate the 403B FUNCTION switch to ON and
RANGE switch to 0.1 volt.
d.
Set Voltmeter 'Calibrator C,·hp-Model 738BR)
OUTPUT SELECTOR to 400Hz rms and 300 volts; the AC Voltmeter reading should be less than 3.5 volts. (If necessary, check
CR1 and CR2).
Figure 5-4.
Performance Check Setup
5-24.
CALIBRATION AND TRACKING.
a.
Disconnectthe 403Bfrom the ac power source.
(The following procedure should be performed with battery operation).
b.
Connect the 403B as shown in Figure 5-4
(Position A).
c.
- - - - - N O T E - - - - - -
The 200K resistor is used only for the input resistance check (Paragraph 5-21).
Rotate the 403B RANGE switch to O. 001 volt.
d.
Set the Voltmeter Calibrator (-hp- Model
738BR) for 0.001 volt rms at 400 Hz.
e.
Preset R33 (loop gain adj) 1/3 clockwise am adjust R29 (meter cal. 1 mV range) for a full-scale imication on the 403B.
f.
Rotate the 403B RANGE switch to O. 1 volt.
Set the Voltmeter Calibrator to O. 1 volt rms at 400 Hz.
g.
Adjust R3 (cal. 0.1 V 400 HZ) for scale indication on the 403B meter.
a fullh.
Rotate 403B RANGE switch to 30.0 volts and set Voltmeter Calibrator to 30.0 volts at 400
Hz rms.
j.
Adjust R4 (see Figure 5-2) for a full-scale indication on the 403B meter.
k.
Check calibration on the 0.003, 0.01, and
0.03 volt ranges; accuracy should be within
±2.0% of full scale on all ranges.
m.
Set Voltmeter Calibrator for 1. 0 volt. Rotate
403B FUNCTION switch to 1. 0 volt.
n.
Check 403B meter tracking at O. 1 volt increments.
Variation should be less than ±2% of full scale.
5-4
Model403B section V
4038-8-1324
TEST OSCILLATOR hp652A hp403B
AC VOLTMETER
50n FEEO-THRU
TERMINATION hpll048B
Figure 5-5.
Frequency Response setup
5-25.
mGH FREQUENCY RESPONSE.
a.
Connect the 403B as shown in Figure 5-5.
b.
set oscillator (-hp- Model 652A) to 1 mV rms range at 400 Hz.
c.
Adjust OUTPUT AMPLITUDE control for reading of 0.94 mV on the 403B.
d.
Place OUTPUT MONITOR switch to expand and adjust REF SET for a convenient reference point on 652A meter.
e.
Vary the 652A frequency dial between 1 MHz and 3 MHz keeping the same reference point as in step d. Adjust C16 (cal. 1 mV 2 MHz) for O. 94 m V ±5% frequency response between
1 and 2 MHz with a gradual roll-off between 2 and 3 MHz.
If C16 does not have sufticient range, the value of C15 should be cIlanged
(range of C15 is from 100 pF to 160 pF).
f.
Adjust 652A Oscillator for a 300 kHz output frequency and same output level as in step d.
g.
Rotate 652A frequency dial between 300 kHz and 1 MHz while maintaining same reference on meter. The 403B meter should indicate
0.9 mV ±O. 02 mV (±2%).
If necessary, adjust R33 for a flat response.
- - - - - N O T E - - - - - -
Repeat Paragraphs 5-24 and 5-25 if R33 is adjusted.
h.
Rotate the 403B RANGE switch to O.
1.
j.
set the 652A Oscillator for 400 Hz and O. 1
RANGE.
k.
Adjust the OUTPUT AMPLITUDE control for an indication of 0.96 of full scale on the 403B meter.
m.
Readjust the 652A OUTPUT AMPLITUDE control for a convenient reference on the
652A meter.
n.
Change the 652A frequency to 300 kHz while maintaining reference on meter.
p.
Adjust C2 for a O. 96 of full scale reading on the 403B meter.
5-26.
10 VOLT RESPONSE CALIBRATION.
a.
Turn the 403B RANGE switch to the 10 V
RANGE.
f.
Connect Wide Range Oscillator (-hp- Model
200CD) to the 403B INPUT. set the Oscillator frequency to 400 Hz.
c.
Connect anAC Voltmeter (-hp-ModeI400E/EL) to the Oscillator OUTPUT.
d.
Adjust the Oscillator AMPLITUDE for 9.40
volt indication on the 403B meter.
Record the AC Voltmeter (-hp- Model 400E/EL) reading.
e.
set the oscillator frequency to 300 kHz. Adjust the Oscillator AMPLITUDE control until the AC Voltmeter indicates the reference level recorded in step d. The 403B should indicate 9.2 to 9.6 volts.
- - - - - N O T E - - - - - -
The AC Voltmeter used in this procedure should have been recently calibrated and have a known frequency response from 400 Hz to at least 300 kHz.
If there is a variation in response between 400 Hz and 300 kHz, this should be considered when adjusting the 403B.
Adjust C3 for a403B meter indication of 10.0
volts.
g.
Repeat Paragraph 5-25, steps h thru p, adjusting C2 and C3 for optimum performance between 0.1 volt range at 300 kHz and 10 volt range at 300 kHz.
5-27.
TROUBLESHOOTING.
5-28.
To assist introubleshooting, Tables 5-3 and5-4 are included in this section of the manual.
Table 5-3,
Troubleshooting, is used for evaluating problems that may be encountered and easily recognized by the operator, and therefore consists mainly of frontpanel inlications.
Tables 5-3 and 5-4. Test Procedure Troubleshooting, is for the technician to localize areas of trouble encountered while testing the
Mode1403B.
- - - - - NOTE - - - - - -
When replacing any crystal diodes or transistors in the Model 403B, refer to Paragraph 5-33 and Table
5-5.
5-5
section V
Table 5-3.
Troubleshooting
SYMPTOM
No response to input
CAUSE
Fuse F 1 blown
Batteries low
Shorted transistor
CR1 or CR2 shorted
Open contacts in range switch
Low reading on Batt.
test
Noise indication on known quiet source
Recharge Batteries
CR1 or CR2 noisy
Noisy transistors
(usually Q1 or Q2)
CR3 or CR4
Meter pins when switchDirty contacts in range ing through ranges switch
C7, C12, or C13 leaky
Meter pulsates at frequencies below 15 Hz
C17, 18, 20 open or leaky
Meter calibration off on Resistors or capacitors ranges above 0.03
bad in range switch
Meter calibration off on Resistors bad in interranges below 0.1
mediate attenuator
Dirty contacts in range switch
Battery will not hold charge
CR10 shorted
Shorted cell in battery
Battery charge inoperaQ7, CR5, CR6, CR7, tive CR8, CR9, C21
Switch on 230V position when using 115V
./
0.001, 0.1, 10 cali~raReplace C13 tion okay but all other ranges out of calibration at 400 Hz
If all ranges on 400 Hz calibration check out okay except for one or two ranges and the stick resistors check okay
ChangeQ3
At 3 volts 2 MHz meter Shorten leads on R18.
reads high If this doesn't fix problem, replace R18
No adjustment on chargCheck for solder splashes ing current on backside of R39
5-29.
REPAIR.
5-30.
CABINET REMOVAL.
a.
Top Cover: remove the single screw wtich holds the cover to the rear panel and slide the cover toward the rear.
5-6
Model403B b.
Bottom Cover: remove the flat head screw holding the cover to the rear panel and slide the cover toward the rear.
The bail must be up.
c.
Side Covers: remove the flat head screws which hold the cover to the side casting of the instrument.
Table 5-4.
Test Procedure Troubleshooting
SYMPTOM
R29 will not adjust for full scale indication
CAUSE
CR1, 2
CR3, 4 bad
Q1 through Q6 bad
Noise (403B input termUsually Q1 or Q2 noisy inated with a shielded lOOK resistor)
Input resistance out of specs
Q1 or Q2 bad
C9, C10, Cll, R6
Meter does not track properly
1) Meter reads conCR3, CR4 bad sistently above or below R35 wrong value all meter divisions
2) Meter reads above some but below other divisions
Diodes CR3, CR4 bad
Meter M1 bad
Low frequency response C1, 7, 12, 13, 18-20 bad
400D reads more than
1. 5 volts on overload
CR1 or CR2 bad
Excessive Charging Rate Bad CR9, Q7
R39 No Effect
5-31.
SERVICING ETCHED CIRCUIT BOARDS.
5-32.
One double-sided and two single-sided circuit boards are used in the Model 403B.
Follow these general rules when servicing these boards.
a.
Do not apply excessive heat to the conductor or component being soldered.
b.
Use a toothpick or wooden splinter to clean holes before inserting new components.
c.
To remove a damaged component, clip leads near component; then apply heat and remove component lead with a straight upward motion.
d.
To insure good connection between the eyelet and conductor, solder from the conductor side.
Mode1403B Section V
5-33.
TRANSISTOR REP~CEMENT.
5-34.
Transistors can be damaged by excessive heat.
When replacing transistors on the Mode1403B printed circuit board, use a heat sink, such as an alligator clip, and use a low wattage soldering iron. Refer to Table 5-5 for any adjustments after replacement.
5-35.
FUNCTION SWITCH REPAIR.
5-36.
Figure 5-7 gives cabling detail on Model 403B
FUNCTION switch.
5-37.
nut.
Q5
Q6
Q7
FLUORESCENT INDICATOR DECAL.
5-38.
If the FUNCTION switch is removed for any reason, the fluorescent indicator decal will have to be replaced. This decal has a special adhesive on the back that holds firmly against the FUNCTION switch
To assure positive contact, proceed as follows:
REFERENCE
Q1, 2
Q3, 4
CR1, 2
CR3, 4
CR5,
CR9
CR10 a.
Moisten the back of the decal with a piece of tissue soaked in xylene
am
allow a few minutes for the adhesive to soften.
b.
Place the decal over the FUNCTION switch shaft, adhesive side down.
Position the black area directly over the OFF line on the
Model 403B panel
am
press the decal firmly against the FUNCTION switch nut.
c.
Slide a bushing or nut over the shaft so that it will hold the decal in contact with the
FUNCTION switch nut, and allow about 20 minutes for the adhesive to dry.
d.
Remove the bushing or nut used for weighting and install the small spacer and FUNC-
TION switch knob.
Table 5-5. Transistor Replacement
FUNCTION
CHE CKS OR ADJUSTMENTS
REQUIRED
Q1 and Q2 work together to provide a high input impedance
(Emitter Follower)
Check input impedance
Readjust R29
Check noise
Amplifier (Common emitter)
Amplifier (Common collector)
Amplifies signal (Common base)
Protects Q1 from overload
Readjust R33
Readjust R33
Readjust R33
Meter Diodes
Recheck overload characteristic s
Check noise
Readjust R29
Readjust R45
Rectifier Diodes
Zener Diode
Isolation Diode
Charging Current Regulation
Check battery charge current
Readjust R39
Check battery charge current
Readjust R39
REFER TO
PARAGRAPH
5-17 steps a thru d
5-24 steps athru e
5-16 steps a and b
5-25 steps a thru g
5-25 steps a thru g
5-25 steps a thru g
5-23 steps a thru d
5-16 steps a and b
5-24 steps a thru e
5-20 steps e thru g
5-20 steps a and b
5-20 steps a and b
5-20 steps a and b
5-20 steps a and b
5-7
Section V
A3C5-----,
.---------2Q,
Model403B
. - - - - - - - - - , B A R E . - - - - - . .
RED
T~~~~~L-8
SIA(F)
'-------:~
SIA(R)
; - - - - - - - - - - - - - , - t - - , B A R E , - - - - - - - - - - - - - - - - . . ,
~--A3C2.RI
F I - - - - - ,
SIB(F)
,Jr-----~
A3R4.R5------~
SIB(R)
5-8
403B-C-1316
A
51
IRANGEI
SIC(F)
B c o
- - - - A I C I 3
SID(R)
SIA(F)7
SWITCH~
\
DESIGNATOR SECTION
~CONrACT
FRONT SIDE NUMBER
1.
WAFER FRONTS (F) VIEWED FROM BOTTOM FRONT OF INSTRUMENT.
2.
WAFER REARS (R) VIEWED FROM BOTTOM REAR OF INSTRUMENT.
3.
SlC(Rl AND SlD(F) HAVE NO SWITCHING AND ARE NOT SHOWN.
4.
SWITCH SHOWN IN 1.001/ -60
I
POSITION.
5.
CIRCLED NUMBERS
G)
INDICATE CONTACT POSITIONS EVEN
THOUGH SOME POSITIONS MAY BE BLANK.
Figure 5-6.
Range Switch Detail
SCHEMATIC NOTES
1.
RANGE SWITCH SHOWN ON 0.001 VOLT RANGE.
2.
FUNCTION SWITCH SHOWN IN THE OFF POSITION.
3.
SWITCH CONNECTIONS ARE SHOWN FOR SCHEMATIC SIMPLICITY
AND MAY NOT BE ACCOMPLISHED IN THE MANNER SHOWN.
5.
4.
ALL DC VOLTAGES ARE TYPICAL VALUES.
A TOLERANCE OF ± 10%
SHOULD BE ALLOWED WHEN MEASURING THESE VOLTAGES.
W
DENOTES CHASSIS GROUND.
W
DENOTES SIGNAL GROUND (FLOATING).
*
OPTIMUM VALUE SELECTED AT FACTORY. AVERAGE VALUE
IS SHOWN AND IN SOME CASES THE PART MAY BE OMITTED.
6.
RESISTOR VALUES ARE IN OHMS AND CAPACITANCE VALUES ARE
IN MICROFARADS UNLESS OTHERWISE DESIGNATeD.
7.
- - - - - - - DENOTES ASSEMBLY.
- - - - - - - - - - - DENOTES MAIN SIGNAL PATH.
8. c==::JDENOTES FRONT PANEL MARKING.
~
DENOTES SCREWDRIVER ADJUST.
9.
~ DENOTES WIRE COLOR: COLOR CODE SAME AS RESISTOR
COLOR CODE.
FIRST NUMBER IDENTIFIES BASE COLOR.
SECOND NUMBER IDENTIFIES WIDER STRIP, THIRD
NUMBER IDENTIFIES NARROWER STRIP.
(e.
g.
~
= WHITE. RED. YELLOW.)
A 2 T 1 - I - - - J ! t - - - - . . .
~----~----J2
-13V TO A I - 2
BT4(~)----L
A2R41----.:L
- - - - - , B A R E , - - - - - . /
S2B(F)
""",__ --~---R45 b - - -
A 2CRIO b - - - B T I ( + )
S-+13V TO AI
52
I
FUNCTION
I
M I ( - ) - - - - J -
' - - - - - - ,BAR!:., - - f - - - - - '
4038-8-1317
S2A(F)
- - - - - - - - & , - - -
BT3H
~--6.5V
TO AI
~---A2R44
~----
~----MI(+)
(F) (F)
A B
1.
SWITCH WAFERS VIEWED FROM BOTTOM
FRONT OF INSTRUMENT.
2.
SWITCH SHOWN IN ~ POSITION.
3.
CIRCLED NUMBERS G) INDICATE CONTACT
POSITIONS EVEN THOUGH SOME POSITIONS
MAY BE BLANK.
Figure 5-7.
Function Switch Detail
Section V Mode1403B
IINPUTI
::-....;J..
I_ _
~
!
~ :r~
I I-Q,
SHORTING
STRAP
~
[i]~11
W
I
RANGE I o
:.°6°ci ill :?~5
Q)
.:~l>
@ :~~
® -~o
® ':I~ o
6
®
+~o
®
~~o
@ 13<&
®
~~<&
@ ;~g
IRANGEI
P/O 51
0
~I~R-(~A~
- - - - - - - - - - - - - - - - - -
---------1 oill o(i)
I
I o@
-13V o® o@
0
0 o®
®
@
I R7
34K
R8
82K
C9
200
-13V
+
AC FEEDBACK
~
(see Para. 4-9)
1
®
®
I
RANGE
I
I
R9
150K
C7 RII
-13V
I
I
CRi
~to0liil,-C>OP/
.....
FI
0-S<lI~.1/Y'16
AMP l 4. . . . .I.,..K"-........
--l F
3
. .
AR;IT;N~~3;-~
-
---'1 ill
~ ~I
R:
82M 4700
3
I
I
I '
2
RIO
2M
I
4
AC FEEDBACK
(see Para. 4-10).
CIO
10
r--------l
I
I
02
. I,.,
..=....:..:I
. . .
---"!~
CI2
200 3419
RI8
I
I
RI9
1081
R2/
341.9
I
RI7
1080
R20
634
®
Q)
I
I
R 4
2 K
I ~b1
~I
I , Ht--~+-i
-13V
R22
920 @
I
I
®o
~/
C 6 * r :
22PF~ ill
0
Q)o
@o
®o
@o
0
0
IRANGEI
0P/0 51
CAL .IV
300KC
~
R2
19./K
I ~
0
.
CAL.,
300t<t:
~I
\V~o
I
@
Tg~OPF ~~ ~~
W ~~5 ~~~
~
T
W
i
2bPF
I
I
@
®
I
~--------~
FROM 97
PWR SOURCE =,:,J1(f-- -13V
VIA 52 {
I
I
I .,,/ASSEMBLY
~
MTO SCREWS
921
(f--+13V
961 T~6=J I
I
I
I
I
I
R23
158.1
'\~
~
~~
0
®
&La
~
@
-<J
@
I
I
I l
I
I
I
CAL IMV
2MC
~
_ _
~
NEG DC
!FEEDBACf<
-13V
R46
4700
-13V
3
R34
8200
R35
390
~03K6
R42
220
18
FUNCTION
I
FUNCTION
I
0
-_-~~}....:6=3::..---O-...L.,~:!.-t-
.......
~
~
~
._ .. -
;U;
~M;L~
(403B-65;) - - - - - - - - - - - - - - - -
R44
220K
BATT
TEST CAL R45
~
2 5 0 K j
I
I
IFUNCTIONI
P/O S2B(F)
. . . - - - - : : 0 - - 0
0 ill
BATT.
OFF o TEST~ON
(j)
12."."".} ~
,-
: 00
P/O S2A(F)
_---J..---
0
: ill (ft
Q)
I
I
I
I
I
I IQ>
WI
J2 oill
Q)
21
A2R38
33K
R47
150
R37
300
07
I
I
I
I
1--~20-~
7
II
~
R4" ill
3900
BTl _
6.5V
Q) 1 c;=-
+13V
BT2 _
6.5vl~
Bn _
6.5V-
I o o ill
16
(i)
~-6.5V
3 o
403B-E-131~B l - - - - - - - -......- -......
- - - - - - - - - - . ; . - 4 o J l
COPYRIGHT 1962 BY HEWLETT -PACKARD COMPANY l
R41
33K
~-------j
I
I
+
MI
Figure 5-8.
Schermtic Diagram
5-9/5-10
Hewlett-Packard Model 403B
Transistorized AC Voltmeter serial No.
DESCRIPTION l.
CALIBRATION
A.
RANGE ACCURACY
300V
100V
30V
10V
3V
1V
0.3V
0.1 V
0.03V
0.01 V
0.003V
0.001 V
2.
HIGH FREQUENCY RESPONSE
A.
0.001 RANGE
500 kHz
1 MHz
2 MHz
B.
0.003 RANGE
500 kHz
1 MHz
2 MHz
C.
0.01 RANGE
500 kHz
1 MHz
2 MHz
D.
0.03 RANGE
500 kHz
1 MHz
2 MHz
E.
0.1 RANGE
500 kHz
1 MHz
2 MHz
F.
3 V RANGE
500 kHz
1 MHz
2 MHz
G.
10 V RANGE
300 kHz
PERFORMANCE CHECK TEST CARD
Tests Performed by
Date
PAR.
O.
CHECK
Full scale
Full scale
Full scale
Full scale
±
±
±
±
2%
2%
Full scale ± 2%
Full scale ± 2%
2%
Full scale ± 2%
Full scale ± 2%
Full scale ± 2%
Full scale ± 2%
Full scale ± 2%
Full scale ± 2%
2%
0.94 Full scale
0.94 Full scale
0.94 Full scale
0.94 Full scale ± 2%
0.94 Full scale ± 2%
0.94 Full scale ± 5%
0.94 Full scale
O.
94 Full scale
94 Full scale
0.94 Full scale
±
±
±
±
2%
2%
5%
2%
O.
94 Full scale ± 2%
0.94 Full scale ± 5%
0.94 Full scale
0.94 Full scale ± 2%
O.
94 Full scale ± 5%
0.94 Full scale
±
±
2%
2%
O.
94 Full scale ± 2%
O.
94 Full scale ± 5%
0.94 Full scale
±
±
±
±
2%
2%
5%
2%
3.
LOW FREQUENCY RESPONSE
A.
0.001 RANGE
10 cps
B.
0.003 RANGE
10 cps
O.
O.
94 Full scale
94 Full scale
±
±
2%
2%
Hewlett-Packard Model 403B
Transistorized AC Voltmeter
Serial No.
PERFORMANCE CHECK TEST CARD (Cont'd)
Tests performed by
Date
PAR.
DESCRIPTION
3.
LOW FREQUENCY RESPONSE (Cont'd)
C.
0.01 RANGE
10 Hz
D.
0.03 RANGE
10 Hz
E.
O~ 1 RANGE
10 cps
F.
3 RANGE
10 cps
4.
NOISE CHECK
A.
BATTERY OPERATION
B.
AC OPERATION
5.
INPUT RESISTANCE
CHECK
0.94 Full scale ± 2%
O.
94 Full scale ± 2%
0.94 Full scale ± 2%
O.
<
<
94 Full scale ± 2%
3% Deflection
8% Deflection
9.0 to 9.5 mV
Model403B section VI
SECTION
REPLACEABLE
VI
PARTS
6-1.
INTRODUCTION.
6-2.
This section contains information for ordering replacement parts.
Table 6-1 lists parts in. alphameric order of their reference designators and indicates the description, -hp- part number of each part, together with any applicable notes, and provides the following: a.
Total quantity used in the instrument (TQ column).
The total quantity of a part is given the first time the part nUmber appears.
b.
Description of the part. (see list of abbreviations below.) c.
Typical manufacturer of the part in a fivedigit code.
(see Appendix A for list of manufacturers. ) d.
Manufacturer's part number.
6-3.
Miscellaneous parts are listed at the end of
Table 6-1.
6-4.
ORDERING INFORMATION.
6-5.
To obtain replacement parts, address order or inquiry to your local Hewlett-Packard Field Office.
(see Appendix B for list of office locations.) Identify parts by their Hewlett-Packard part numbers. Include instrument model and serial numbers.
6-6.
NON-LISTED PARTS.
6-7.
To obtain a part that is not listed, include: a.
Instrument model number.
b.
Instrument serial number.
c.
Description of the part.
d.
Function and location of the part.
Ag
AI
.sil_
eluminum
A . • . . . . . . . . . . . • . . . . . . . . . . . . . • ernpefels)
Au . . . . . . . . . . . . • . . . . . . . . . . . . . . . . . . gold
C . . . . . . . . . . . . . . . . . . . . • . . . . . . . c:epecitor cer ...................•........ .c...
.."ic coef coefficient corn • • • • • • • • • • • • • • • • • • • • • • • • • • • COf'nlT'On comp ........••......•...... composition conn connection dep
OPOT
OPST deposi ted double-pole double·throw double-pole sinllle·throw elect electrolytic encep .•...•..•...•.•••...•• encepaJleted
F .f.edls)
FET .....••......... fiald effect trensistor fxd . . . . . . . . • . . . . . . . . . . . . . . . . . . . . .fixed
GeAs gellium ._ids
GHz . . . . . . • . . . . . . . .
gigehertz 10+9 hertz gd . . . . . . . . • . . . . . . . . . . . . . . . . • .
~.d(edl
Ge . . . . . . . . • . . . . . . . . . . . . . . . . . gtrmenium end ground(ed)
H . . . . . . . . . . . . . . . . • . . . . . . . . . . .
henry(ies)
Hg mercury
ABBREVIATIONS
Hz . . . . . . . . . . . ..
hertz (cyele(s) per oecond) NPO
10 impg inside di t .
ns impnlllneted nsr
(..-0 negetive positive ..-0
...".,.eture coefficient) nenooecondls) - 10-9 seconds not leJ*"etltly replec:eeble l incd .incendncent
ins . . . . . . . . . . • . . . . . . . . . . . . . in..letion(ed) n kn kilohm(s) - 10+3 ohms ohm(s)
~::::::::::::::::
:
:~.~:u~=~: kHz kilohertz - 10+3 hertz p •.••..••••...•..•••••••.•••••••• peek inductor lin . . . . . . . . . • . . . . . . . . . . . . . . . . line.
teper log ....................• logerithmic teper
~:::::::::::::::::::::::.~=:~ pF picof.ed(s) 10-12 f.eds
piv . . • . . . . . . . . . . . . . . . .
peek in_ _ voltege plo pert 01 rnA
MHz
Mn met tim mlr milliempere(s) - 10-3 _ r e s megehertz - 10+6 hertz megohm(s)· 10+6 ohms
met"
film menufecturer rns •...•..•..••••.•........•.
millisecond pol . • . • • • . • . • . • • • • • • . . • • • . . .
poly position(s) polystyrene pot . . . . . . . . . . . • . . . . . . . . . . potentiometer
Pi> peek·to-peek ppm prec peru per million precision (tempereture coeffie(lt, mtg : '.' ~~nting mV . . . . . . • . . . . . . . . mlillvolt(s) - 10 volts
J.IF ......•.......•.•.....•• microf.ed(s)
/.II
IN microsecond(s) microvolt(s) - 10-6 volts
R
Rh long term stebility end/or toler....,.) resistor rhodium my Myl.® rrTlS. • . . . . . . . • . . . . . • . . . . • root.",.....-squ• • rot rotary nA
Nt::
Ne nenoempere(s)-10-9 emperes normelly closed s. ............................• neon oect
Ienium section (s)
NO . . . . • . . . . . . . . • . . . . . . . . normelly open Si . . . . . . . . . . . • . . . . . . • . . . . . . . . . . .silicon
II
SPOT
SPST
' II ide single-pole double-throw single-pole single-throw
Te
TC tentelum tempereture coefficient
Ti02 titenium dioxide tog . . . . . . . . . • . . . . . . . . . . . . . . . . . . . toggle to! .........•...............•.
toler....,.
trim , trim"....
TSTR trensistor
V
VIICW • • • • •
_
¥dew volt(s) alterneting current working volteglt v.ieble
direct current working volteglt w wett(s)
wI .......... .•...................
with wiv ..............• Working in_. volt.
wlo without
YtIW • • • • • • • • • • • • • • • • • • • • • • • • • wirewoun<:t
. . .
.....optimum velue .I«ted at fectory,
_ . velue shown (pert mey be omitted) no stenderd type number _gned
"ected or speciel type
®
Dupont de Nemours
DECIMAL MULTIPLIERS
....elix
Symbols Multiplier ....elix
Symbols Multiplier
A
8 mbly motor
BT . . . . . . . . . . . . . . . . . • . . . . . . . . . . .
bettery
C cepecitor
CR . . . . . . • . . . . . . . . . . . . . • . . . . . . . . diode
Ol deley line
OS . . . . . . . • . . . . . . . . . . . . . . . . . . . . . . . Iemp
E
F mioc alectronic pert
','" .•........•.............. fu.
tere gige rnege kilo hecto deke deci
T
G
M or Meg
Kor k h da d
1012
109
106
103
102
10
10.1
centi milli micro pico femto etto
IJ.
10-2
10.3
10-6
10.9
10.12
10-15
10.18
Fl
HR
DESIGNATORS filter a heater OCR transistor transistor-<liode
IC integreted circuit R resistOf
J .....••••....•.•................
jeck RT . . . . . . . . . . . . . • . . . . . . . . . . . thermistor
K . . . . . . • . . . . . . . . . . . . . . . . . . . . . . . . . relay S l . . . . . . . . . . . . . . • . . . . . . . . . . . . . . inductor T
M . . . . . . . . . . . . . . • . . . . . . . . . . . . . . . meter T8
.-itch transformer terminer bo.d
MP ....••.....•.•....... mechenicel pert TC . . . . . . • . . . . . . . . . . . . . . . . . thermocouple
P plug TP .....••......••............ test point
STD-8·2734
TS
U
V
W
X t.minel
strip microcircuit vecuum tube, neon bulb.photocell, atc.
cellIe lOCket
XOS Iempholder
XF . . . . . . . . . • . . . . . . . . . . . . . . . fu.holder
Y ...•..••.•.••.......•.•.•.•.•• crystel
Z .
network
6-1
section VI
REFERENCE
DESIGNATOR
Al
-hp-
PART NO.
403B-65A
A2 403B-65B
A3 403B-65C
BTl, 2, 3, 4
Cl
C2, C3
C4
C5
*C6
C7
C8
C9
CI0
Cll
,
C12
C13
C14
*C15
C16
C17, C18
C19
C20
C21
CR1, CR2
CR3, CR4
CR5 thru
CR8
CR9
CRI0
DSl
Fl
Jl
J2
Ml
Ql,Q2, Q3, Q5
Q4, Q6
1510-0008
1510-0009
5060-0626
0340-0090
1251-2357
1120-0315
1120-0316
1853-0020
1854-0071
1420-0015
0170-0033
01~0-0003
0140-0151
0140-0178
0140-0145
0180~100
0160-0205
0180-0060
0180-0059
0180-0064
0180-0104
0180-0063
0180-0039
0140-0176
0130-0017
0180-0058
0180-0033
0180-0150
0180-0149
...
1901-0044
1901~535
1901-0025
1902-0074
1901-0025
1450-0419
2110-0011
Model403B
Table 6-1 Replaceable Parts (Cont'd)
TQ DESCRIPTION MFR.
MFR. PART NO.
Assembly, printed circuit: includes,
C7 thru C20 Ql thru Q6
CRl thru CR4 R6 thru R16
R24 thru R36
R40
R46
1
2
1
1
1
Assembly, resistor board: includes,
C21 R37 thru R39
CR5 thru CRI0 R41
Tl
Q7
R44
R45
R47
Assembly, resistor board: includes,
C2 thru C5 Rl thru R5
4 Battery, Nickel Cadmium 6. 5 Vnom. 225 mAh 88220 6.0V/225B
C: fxd O. 18 /-LF ±10% 600 vdcw
C: var cer 1. 5-7 pF ±10% 500 vdcw
C: fxd mica 820 pF ±2% 300 vdcw
C: fxd mica 560 pF ±2% 300 vdcw
C: fxd mica 22 pF ±5% 500 vdcw
1 C: fxd elect 4.
7 /-L F .±10% 35 vdcw
1 C: fxd mica 10 pF ±5% 500 vdcw
1 C: fxd elect 200 /-LF +100% -10% 3 vdcw
1 C: fxd elect 10 /-L F 10 vdcw
1 C: fxd elect 35 /-LF -+,100% -10% 3 vdcw
1 C: fxd elect 200 /-LF 15 vdcw
1 C: fxd elect 500 /-LF +100% -10% 3 vdcw
1 C: fxd elect 100 /-L F 12 vdcw
1 C: fxd mica 100 pF ±2% 300 vdcw
1 C: var cer 8-50 pF 500 vdcw
2 C: fxd elect 50 /-LF +100% -10% 25 vdcw
1 C: fxd elect 50 /-L F 6 vdcw
1 C: fxd elect 1200 /-LF 10 vdcw
1 C: fxd elect 65 /-L F 60 vdcw
09134 Type 27
72982 503-000-COPO-I0R
04062 DM15F821G
04062 DM15F561G
04062 DM15C220J
56289 150D475X9035B2-DY~
04062 DM15C220J
56289 30D116Al
56289 30D182Al
56289 30D122Al
56289
56289
56289
04062
72982
56289
56289
56289
56289
30D174Al
30D120Al
30D154Al
557-019-U2 P034R
30D186Al
30D133Al
Type 340
Type 30D
2
2
5
Diode: Si 20 mA/+IV 10 mA/-I0V/50 wiv
2 pF
Diode: hybrid hot carrier
Diode: Si 100 rnA 100 wiv 2 pF
1 Diode: breakdown 7.
15V ±5% 400 mW
Diode: Si 100 rnA 100 wiv 2 pF
73293 HD
-hp':'
03877 SG-817
07910 CE 35664
03877 SG-817
72765 599-237 white
1 Indicator: neon
1 Fuse, 1/16 amp 250 V maximum 5.4 ohm
3 Terminals: three female
1 Assembly binding post: red
1 Assembly binding post: black
1 Assembly binding post: black with strap
1 Insulator: B. P. double keyed
1 Connector: ac power
75915 #312.062
-hp-
-hp-
-hp-
-hp-
82389 EAC-301
4
2
1 Meter: 0-100 /-LA dc
1 Meter: 0-100 /-LA dc DB Scale (option 01)
TSTR: Si PNP
TSTR: Si NPN
-hp-
-hp-
01295 SKA1123
01295 SKA1124
6-2
Model403B
REFERENCE
DESIGNATOR
-hp-
PART NO.
S3
1853-0016
0727-0287
0727-0443
2100-0390
0727-0056
0687-8251
0698-4493
0758-0022
0687-1541
0687-4721
0693-1021
0687-5621
0687-1221
0687-1531
0687-6831
0687-6821
0727-0103
403B-26A
403B-26B
0727-0084
403B-26C
0727-0096
403B-26D
0758-0074
0758-0076
0758-0073
0687-1031
0727-0017
2100-0240
0727-0050
403A-26G
0687-1031
2100-0154
0758-0048
0687-3911
0687-1031
0686-3015
0687-3331
2100-0391
0687-5621
0687-3331
0687-2211
0687-3921
0687-2241
2100-0144
0687-4721
0758-0007
403B-19W
403B-19A
7123-0101
3101-1234
Q7
Rl
R2
R3, R4
R5
*R6
R7
R8
R9
RIO
R11
R12
R13
R14
R15
R16
R37
R38
R39
R40
R41
R42
R43
R44
R45
R46
R47
Sl
S2
R17
R18
R19
R20
R21
R22
R23
R24
R25
R26
R27
R28
R29
R30
R31
R32
R33
R34
R35
R36
Section VI
Table 6-1.
Replaceable Parts (Cont'd)
TQ DESCRIPTION MFR.
MFR. PART NO.
1 TSTR: PNP 2N3638 01295 2N3638
1 R: fxd comp 2M ±1% 1/2 W
1 R: fxd comp 19. lkn ±1% 1/2 W
1 R: var comp dual 2k and 6k ohms 1-1/4 W
1 R: fxd carbon flm 216 ohms ±1/2% 1/2 W
1 R: fxd comp 8. 2M ±10% 1/2 W
19701 DC 1/2 CR5
19701 DC 1/2 CR5
71590 Series Type 73-2
19701 DC 1/2 AR5
01121 EB 825J.
1 R: fxd flm 34 kn ± 1% 1/8 W
1 R: fxd comp 82kn ±5% 1/2 W
1 R: fxd comp 150kn ±10% 1/2 W
2 R: fxd comp 4. 7kn ±10% 1/2 W
1 R: fxd comp lkn ±10% 1/2 W
2 R: fxd comp 5. 6kn ±10% 1/2 W
1 R: fxd comp 1. 2kn ±10% 1/2 W
1 R: fxd comp 15kn ±10% 1/2 W
1 R: fxd comp 68kn ±10% 1/2 W
1 R: fxd comp 6. 8kn ±10% 1/2 W
1 R: fxd carbon flm 1080n ±1% 1/2 W
1 R: fxd ww 3. 41kn ±O. 2% 1/2 W
1 R: fxd ww 1. 081kn ±O. 2% 1/2 W
1 R: fxd carbon flm 634n ±1% 1/2 W
1 R: fxd ww 341. 9n ±O. 2% 1/2 W
1 R: fxd carbon flm 920n ±1% 1/2 W
1 R: fxd ww 158. In ±O. 2% 1/2 W
1 R: fxd met oxide 27kn ±5% 1/2 W
1 R: fxd met oxide 68kn ±5% 1/2 W
1 R: fxd met oxide 24kn ±5% 1/2 W
2 R: fxd comp 10kn ±10% 1/2 W
1 R: fxd carbon flm 37. 35n ±1/2% 1/2 W
1 R: var ww 50n ±20% 1 W
1 R: fxd carbon flm 180n ±1% 1/2 W
1 R: fxd ww 2 sect 30n
91637 CMF-l/I0-32 T-l
07115 C20
01121 EB 1541
01121 EB4721
01121 HB 1021
01121 EB 5621
01121 EB 1221
01121 EB 1531
01121 EB 6831
01121 EB 6821
19701 DC 1/2 CR5
-hp-
-hp-
19701 DC 1/2 CR5
-hp-
19701 DC 1/2 CR5
-hp-
07115 C5
07115 C5
07115 C5
01121 EB 1031
19701 DC 1/2 CR5
11236 Series 110
19701 DC 1/2 CR5
-hp-
R: fxd comp 10kn ±10% 1/2 W
1 R: var comp lkn ±30% 3/10 W
1 R: fxd met oxide 8. 2kn :lr5% 1/2 W
1 R: fxd comp 390n ±10% 1/2 W
R: fxd comp 10kn ±10% 1/2 W
1 R: fxd comp 300n ±5% 1/2 W
2 R: fxd comp 33kn ±10% 1/2 W
1 R: var ww lkn ±20% 1. 25 W
R: fxd comp 5. 6kn ±10% 1/2 W
R: fxd comp 33kn ±10% 1/2 W
01121 EB 1031
11237 UPE70 C5
07115 C5
01121 EB 3911
01121 EB 1031
01121 EB 3015
01121 EB 3331
11236 Series 110
01121 EB 5621
01121 EB 3331
1 R: fxd comp 220n ±10% 1/2 W
1 R: fxd comp 3. 9kn ±10% 1/2 W
1 R: fxd comp 220kn ±10% 1/2 W
1 R: var comp 250kn ±30% 0.2 W
R: fxd comp 4. 7kn ±10% 1/2 W
1 R: fxd met oxide 150n ±5% 1/2 W
1 Assembly RANGE switch 3 sect 12 pos includes: C2 thru C6, A3, and R17 thru'R23
1 Assembly FUNCTION switch 2 sect 3 pos includes: R42 and R43
1 Washer fluorescent irrlicator for use with
Function Switch Knob
1 Switch-Slide: DPDT 115-230V
01121 EB 2211
01121 EB 3921
01121 EB 2241
11237 UPE70
01121 EB 4721
07115 C5
-hp-
-hp-
-hp-
82389 11A-1242A
6-3
L . . . . -_ _
section VI
REFERE CE
DESIGNATOR
Tl
WI
XFl
-hp-
PART NO.
9100-0172
8120-1348
1400-0008
00403-90008
Model403B
Table 6-1.
Replaceable Parts (Cont'd)
TQ DESCRIPTION MFR.
MFR.
PART NO.
1
1
Transformer
Assembly cable power
98734 6-2249
70903 KHS-7041
1 Holder fuse 1/2" wide 3/16" thick 1-5/8" long 75915 3510-11
MISCELLANEOUS
·1 Operating and service Manual -hp-
..
,-
6-4
Model403B
Section VI
Table 6-2.
Mechanical Parts
Cabinet Parts
CD
0[)::
> >
>
Item
Number
9
10
7
8
5
6
1
2
3
4
~ Stock No.
5060-0702
1490-0031
5040-0700
5060-0727
5020-0700
5000-8563
5060-8547
5000-8564
00403-00202
00403-00203
Description
Frame Assembly
Stand: Tilt
Hinge
Foot Assembly
Spacer
Cover: Side
Cover Assembly: Top
Cover Assembly: Bottom
Panel: Rear
Panel: Front
(See Figure 6-1. )
Note
6-5
Section VI Model403B
Table 6-2.
Mechanical Parts (Cont'd)
44
45
46
47
48
49
23
24
25
26
27
19
20
21
22
15
16
17
18
9
10
11
12
13
14
28
29
30
31
32
33
34
39
40
41
42
43
35
36
37
38
50
51
52
53
54
NUMBER ON
ILLUSTRATION
ON FIGUa,E 6-1
1
2
3
4
5
6
7
8
55
56
57
NAME/DESIGNATOR STOCK NUMBER
Indicator, Neon
Retainer clip
Insulation, vinyl tubing (specify 1" length)
Knob, bar w/ indicator, black
Special washer 3/8 inch OD x 0.26 inch ID
Wafer, fluorescent indicator for use with Function Switch Knob
Panel, front
Meter, 0-100 J1.A
dc (403B) or
Meter, 0-100 J1.A
dc, DB Scale (Option 01)
3/8" - 32 x 1/2" nut, hex.
AC shield
Vertical shield
Bushing, threaded 3/8 - 32
6.32 x 5/16" nut, hex. w/lock
Assembly FUNCTION Switch: 2 sect, 3 pos
Screw 6-32 x 3/8 flat head phillips drive
Screw 6-32 x 3/8 flat head slot drive
Cover 1/3 module 8 inch deep, top
Assembly, printed circuit
Battery holder
Screw, 6-32 x 1-1/2 binding head with lock
Screw 6-32 x 3/8 pan head
Not assigned
Rear panel, 1/3 module, 1/2 recess
Side frame 6 x 8 sub-module
Switch shield
Side cover, 6 x 8, SM
Same as 15
Bottom cover 1/3 module, 8 inch DP
Assembly RANGE Switch: 3 sect, 12 pos
Same as 15
6. 32 x 5/16" nut
#6 split lock, SS same as 16
Spacer No. 6 x 5/16
318" ID x 5/8" OD flat washer
#6 internal lock washer
3/8" internal lock (heavy) washer
#6 solder lug "L"
Tinnerman retainers same as 9
Same as 13
Coupler, shaft 1/4"
#10 solder lug same as 43
#10 internal-external lock washer
#10 internal lock washer
Insulator B. P. double without locating key
Hinge stand, third mod. tilt
Foot assembly, third mod.
Assembly, dial
Insulator, binding post: double keyed
Spacer binding post
Assembly, binding post: black w/strap
Assembly, binding post: red
Assembly, binding post: black
Meter trim, third mod.
Battery, nickel cadmium, 6.25 V nom. 225 mAh
Assembly, resistor board
See DS1, table 6-1
0510-0123
0890-0057
0370-0087
3050-0014
7123-0101
00403-00203
See M1, Table 6-1
2950-0001
403B-6D
403B-6A
1410-0003
2420-0001
See S2, Table 6-1
2370-0013
2370-0002
5060-8547
See A1, Table 6-1
5040-0615
2370-0010
2390-0010
00403-00202
5060-0702
403B-6B
5000-8563
5000-8ij69
See Sl, Table 6-1
2420-0002
2190-0n06
0380-0007
3050-0067
2190-0007
2190-0022
0360~042
0590-0039
5020-0237
0360-0007
2190-0028
2190-0011
0340-0086
5040-0700
1490-0031
5060-0727
403B-99
0340-0090
0400-0084
See J1, Table 6-1
5020-6852
See BTl, 2,3,4, Table 6-1
See A3, Table 6-1
6-6
Model403B Section VI
Figure 6-1.
Exploded V· lew
6-7
Appendix A
CODE LIST OF MANUFACTURERS
The following code numbers are from the Federal Supply Code for Manufacturers Cataloging Handbooks H4-1
(Name to Code) and H4-2 (Code to Name) and their latest supplements. The date of revision and the date of the supplements used appear at the bottom of each page. Alphabetical codes have been arbitrarily assigned to suppliers not appearing in the H4 Handbooks.
Model403B
Cod.
No.
Manufoctur~
Address
00000 U. S. A. Common
00136 MCCoy Electronics
00213 Saee Electronics Corp.
00287 Cemco Inc.
00334 Humidial
00348 Microtron Co., Inc.
00373 Garlock Inc.
Any supplier of U. S.
Mount Holly Sprines, Pa.
Rochester, N. Y.
Danielson, Conn.
Colton, Calif.
Valley Stream, N. Y.
Cherry Hill, N. J.
00656 Aerovox Corp.
00779 Amp. Inc.
New Bedford, Mass.
Harrisbure, Pa.
00781 Amrafl Radio Corp.
Boonton, N. J.
00815 Norlhern Eneineerine Laboratories, Inc.
Burlineton, Wis.
00853 Saneamo Electric Co., Pickens Div.
00866 Goe Eneineerine Co.
00891 Carl L Holmes Corp.
Pickens, S. C.
City of Industry, Cal.
Los Aneeles, Calif.
00929 Microlab Inc.
Livineston, N. J.
01002 General Electric Co., Capacitor Dept.
Hudson Falls, N. Y.
01009
01121
01255
01281
01295
01349
01589
01930
01961
02114
02116
02286
02660
02735
Alden Products Co.
Allen Bradley Co.
Litton Industries, Inc.
TRW Semiconductors, Inc.
Texas Instruments, Inc.,
Transistor Products Div.
The Alliance Mfe. Co.
PacifiC Relays, Inc.
Amerock Corp.
Pulse Eneinetrine Co.
Ferroxcube Corp. of America
Wheelock Siena Is, Inc.
Cole Rubber and Plastics Inc.
Brockton, Mass.
Milwaukee, Wis.
Beverly Hills, Calif.
Amphenol·Borg Electronics Corp.
Lawndale, Calif.
Dallas, Texas
Alliance, Ohio
Van Nuys, Calif.
Rockford, III.
Santa Clara, Calif.
Saueerties, N. Y.
Lone Branch, N. J.
Sunnyvale, Calif.
Chicago, III.
Radio Corp. of America, Semiconductor
02771 and Materials Div.
Vocallne Co. of America, Inc.
Somerville, N. J.
Old Saybrook, Conn.
02777 Hopkins Engineering Co.
03508 G. E. Sem Iconductor Prod.
03705 Apex Machine & Tool Co.
San Fernando, Calif.
Dept.
Syracuse, N. Y.
Dayton, Ohio
03797 Eldema Corp.
03818 Parker Seal Co.
03877
03888
Transitron Electric Corp.
Pyrofllm Resistor Co., Inc.
Compton, Calif.
Los Angeles, Calif.
Wakefield, Mass.
Ce ar Knolls, N.
J.
03954 Singer Co., Diehl Div.
Finderne Plant Sumerville, N. J.
04009 Arrow, Hart and Hegeman Elect. Co.
Hartford, Conn.
04013 Taurus Corp.
04062 Arco Electronic Inc.
04222 HI·Q DiVision of Aerovox
04354 PreCIsion Paper Tube Co.
Lambertville, N. J.
Great Neck, N. Y.
Myrtle Beach, S. C.
Wheeling, III.
04404 Dymec DiVision of Hewlett·Packard Co.
Palo Alto, Calif.
04651 Sylvania Electric Products, Microwave
Device Dlv.
Mountain View, Calif.
04713 Motorola, Inc., Semiconductor Prod. Div.
Phoenix, Arizona
04732 Filtron Co., Inc. Western Div.
Culver City, Calif.
04773 Automatic Electrrc Co.
04796 Sequoia Wire Co.
04811 PreciSion COil Sprine Co.
04870 P. M. Motor Company
04919 Component Mfg. Service Co.
Northlake, III.
Redwood City, Calif.
EI Monte, Calif.
Westchester', III.
W. Brideewater, Mass.
05006 Twentieth Century Plastics, Inc.
Los Angeles, Calif.
05277 Wesllnghouse Electric Corp.
Seml·Conductor Dept.
Younewood, Pa.
Code
No.
Manufacturer Address
05347
05397
UItronix, Inc.
San Mateo, Calif.
Union Carbide Corp., Linde Div., Kemet Dept.
05574
05593
05616
05624
05728
VikIng Ind. Inc.
IlIumltronlc Engineering Co.
Cosmo Plastic
(c/o Electrical Spec. Co.)
Cleveland, Ohio
Canoga Park, Calif.
Sunnyvale, Calif.
Barber Colman Co.
Tiffen Optical Co.
Cleveland, Ohio
Rockford, III.
Roslyn Heights, Long Island, N. Y.
Metro·Tel Corp.
Stewart Engineering Co.
Westbury, N. Y.
Santa Cruz, Calif.
05729
05783
05820
06004
. 06090
06175
06402
06540
Wakefield Engineering Inc.
Wakefield, Mass.
Bassick Co., Div. of Stewart Warner Corp.
Raychem Corp.
Bausch and Lomb Optical Co.
Bridgeport, Conn.
Redwood City. Calif.
Rochester, N. Y.
E. T. A. Products Co. of America
Amalom Electronic Hardware Co., Inc.
Chicago, III.
New Rochelle, N. Y.
Beede Electrical Instrument Co., Inc.
06555
06666 General Devices Co., Inc.
06751 Semcor Div. Components Inc.
06812 Torrington Mfg. Co., West Div.
06980
07088
07126
07137
07138
07149
07233
07256
07261
07263
07322
07387
07397
Penacook, N. H.
Indianapolis, Ind.
Phoenix, Ariz.
Varian Assoc. Eimac Dlv.
Kelvin Electric Co.
Dlgltran Co.
Transistor Electronics Corp.
Van Nuys, Calif.
San Carlos, Calif.
Van Nuys, Calif.
Pasadena, ,Calif.
Minneapolis, Minn.
Westinghouse Electric Corp.
Electronic Tube Div.
Filmohm Corp.
Cinch·Graphik Co.
SilIcon Transistor Corp.
Elmira, N. Y.
New York, N. Y.
City of Industry, Calif.
Carle Place, N. Y.
Avnet Corp.
Fallchlld Camera & Inst. Corp.
Semiconductor Div.
Culver City, Calif.
Mountain View, Calif.
Minnesota Rubber Co.
Blltcher Corp., The
Minneapolis, Minn.
Monterey Park, Calif.
07700
07829
07910
07933
Sylvania Elect. Prod. Inc., Mt. View Operations
Technical Wire Products Inc.
Mountain View, Calif.
Cranlord, N. J.
BodIne Elect. Co.
Continental Device Corp.
Chicago, III.
Hawthorne, Calif.
07980
08145
08289
08358
Raytheon Mfg. Co ..
Semiconductor Dlv.
Hew lell·P ackard Co. ,
Mountain View. Calif.
Boonton Radio Div.
U. S. Engineering Co.
Blinn, Delbert Co.
Rockaway. N.
J.
Los Angeles, Calif.
Pomona, Calif.
Burgess Battery Co.
Niagara Falls, Ontario, Canada
08524
08664
08717
08718
Deutsch Fastener Corp.
Bristol Co., The
Sloan Company
Los Angeles, Calif.
Waterbury, Conn.
Sun Valley, Calif.
ITT Cannon Electric Inc., Phoenix Div.
08727 National Radio Lab. Inc.
08792 CBS ElectronIcs Semiconductor
Operations, Div of C. B. S. Inc.
PhoeniX, Arizona
Paramus, N. J.
Lowell, Mass.
08984 Mel-Rain
09026 Babcock Relays Div.
09134 Texas Capacitor Co.
09145 Tech. Ind. Inc. Atohm Elect.
09250 Eleclro Assemblies, Inc.
09569 Mallory Battery Co. of
Canada, Ltd.
Indianapolis, Ind.
Costa Mesa, Calif.
Houston, Texas
Burbank, Calif.
Chicago, III.
Toronto. Ontario, Canada
10214 General TranSistor Western Corp.
Los Angeles, Calif.
Code
No.
Manufacturer Address
10411 Ti-Tal, Inc.
10646 Carborundum Co.
Berkeley, Calif.
Niagara Falls, N. Y.
11236 CTS of Berne, Inc.
11237 Chicaeo Telephone of California, Inc.
Berne, Ind.
So. Pasadena, Calif.
11242 Bay Slate Electronics Corp.
Wallham, Mass.
11312 Teledyne Inc., Microwave Div.
Palo Alto, Calif.
11314 National Seal
11534 Duncan Electronics Inc.
Downey, Calif.
Costa Mesa, Calif.
11711 General Instrument Corp., Semiconductor
Div., Products Group Newark, N. J.
11717 Imoerral Electronic, Inc.
11870 Melabs, Inc.
12136 Philadelphia Handle Co.
12361 Grove Mfg. Co., Inc.
Buena Park, Calif.
Palo Allo, Calif.
Camden, N.J.
Shady Grove, Pa.
12514 Gullon Ind. Inc. Data System Div.
Albuquerque, N. M.
12697 Clarostat MIg. Co.
12728 Elmar Filler Corp.
12859 Nippon Electrtc Co., Ltd.
12881 Metex Electronics Corp.
12930 Della Semiconductor Inc.
12954 Dickson Electronics Corp.
13103 Thermolloy
13396 Telefunken (GmbH)
Dover, N. H.
W. Haven, Conn.
Tokyo, Japan
Clark, N.J.
Newport Beach, Calif.
Scottsdale, Arizona
Dallas, Texas
Hanover, Germany
13835 Midland-Wright Div. of Pacific Industries, Inc.
Kansas City, Kansas
14099 Sem·Tech Newbury Park, Calif.
14193 Calif. ReSistor Corp.
Santa Monica, Calif.
14298 American Components, Inc.
. Conshohocken, Pa.
14433 ITT Semiconductor, A Dlv. of Int. Telephone
& Telegraph Corp.
West Palm Beach, Fla.
14493 Hewlell·Packard Company
14655 Cornell Dublier Electric Corp.
14614 Corning Glass Works
14752 Electro Cube Inc.
14960 Williams Mfg. Co.
15203 Webster Electronics Co.
15287 SClonlcs Corp.
152ql Adjustable BUShing Co.
15558 Micron Electronics
Loveland, Colo.
Newark, N. J.
Corning, N. Y.
San Gabriel, Calif.
San Jose, Calif.
New York, N. Y.
Northridge. Calif.
N. Hollywood, Calif.
Garden City, Long Island, N. Y.
18324
18476
18486
18583
18612
18873
18911
19315
16758
17109
17474
17675
17745
17870
18042
18083
15566 Amprobe lnst. Corp.
15631 Cabletronlcs
Lynbrook, N. Y.
Costa Mesa, Calif.
15772 Twentieth Century Coil Spring Co.
Santa Clara, Calif.
15801
15818
16037
16179
16352
16688
Fenwal Elect. Inc.
Amelco Inc.
Spruce Pine Mica Co.
Omni-Spectra Inc.
Computer Diode Corp.
Ideal P,ec. Meter Co., Inc.
Framingham. Mass.
Mt. View, Calif.
Spruce Pine, N. C.
Detroit, III.
Lodi, N. J.
De Jur Meter D,v.
Delco RadiO Dlv. of G.M. Corp.
Brooklyn, N. Y.
Kokoma, Ind.
Thermonet,cs Inc.
Tranex Company
HamlIn Metal Products Corp.
Angstrohm Prec. Inc.
McGraw-Edison Co.
Power DeSign Paclfrc Inc.
Canoga Park, Calif.
Mountain View, Calif.
Akron, OhiO
No. Hollywood, Calif.
Manchester, N. H.
Clevlte Corp., Semiconductor Div.
Palo Alto, Calif.
Signetics Corp.
Ty·Car Mfg. Co., Inc.
TRW Elect. Compo Dlv.
Curlis Instrument, Inc.
Vlshay Intertechnology. Inc.
LI. DuPont and Co., Inc.
Durant Mfg. Co.
Palo Alia, Calif.
Sunnyvale, Calif.
Holliston, Mass.
Des Plaines, III.
Mt. Kisco. N. Y.
Malvern, Pa.
Wilmington, Del.
Milwaukee, Wis.
The BendiX Corp., NaVigation & Control Div.
Teterboro, N. J.
00015-45
ReVised: Aprrl, 1968
From: FSC.
H4-1
H4-2
Handbook Supplement.
Dated AUGUST 1966
Dated NOV.
1962
Mode1403B Appendix A
CODE LIST OF MANUFACTURERS (Cont'd)
Code
No.
Manufacturer
Address
19500 Thomas A. Edison Industries, Div. of
McGraw·Edison Co.
West Orange, N. J.
19589 Concoa
19644 LRC Electronics
19701 Electra Mfg. Co.
20183 General Atronics Corp.
Baldwin Park, Calif.
Horseheads, N. Y.
Independence, Kansas
Philadelphia, Pa.
21226 Executone, Inc.
21335 Fafnll Bearing Co., The
Long Island City, N. Y.
New Britain, Conn.
21520 Fansteel Metallurgical Corp.
N. Chicago, III.
23783 Blltish Radio Electronics Ltd.
Washington, D. C.
24455 G. E. Lamp Division
Nela Park, Cleveland, Ohio
24655 General Radio Co.
24681 Memcor Inc., Compo Div.
West Concord, Mass.
Huntington, Ind.
26365 Gries Reproducer Corp.
New Rochelle, N. Y.
26462 Grobet File Co. of America, Inc.
Carlstadt, N. J.
26851 Compac/Hollister Co.
26992 Hamilton Watch Co.
28480 Hewlelt-Packard Co.
28520 Heyman Mfg. Co.
30817 Instrument Specialties Co., Inc.
Hollister, Calif.
Lancaster, Pa.
Palo Alto, Calif.
Kenilworth, N. J.
Little Falls, N.
J.
33173 G. E. Receiving Tube Oept.
35434 Lectrohm Inc.
Owensboro, Ky.
Chicago, III.
36196 Stanwyck Coil Products Ltd.
Hawkesbury, Ontario, Canada
36287 Cunningham, W. H.
& Hill, Ltd.
Toronto Ontario, Canada
37942 P. R. Mallory & Co. Inc.
39543 Mechanical Industries Prod. Co.
Indianapolis, Ind.
40920 Miniature Precision Bearings, Inc.
42190 Muler Co.
43990 C. A. Norgren Co.
Akron, Ohio
Keene, N. H.
Chicago, III.
Englewood, Colo.
44655 Ohmite Mfg. Co.
46384 Penn Eng.
& Mfg. Corp.
Skokie, III.
Doylestown, Pa.
47904 Polaroid Corp.
Cambridge, Mass.
48620 PreciSion Thermometer & Inst. Co.
Southampton, Pa.
49956" Microwave & Power Tube Div.
52090 Rowan Controller Co.
52983 Sanborn Company
54294 Shallcross Mfg. Co.
Waltham, Mass.
Westminster, Md.
Waltham, Mass.
Selma, N. C.
55026 Simpson Electric Co.
55933 Sonotone Corp.
Chicago, III.
Elmsford. N. Y.
55938 Raytheon Co. Commercial Apparatus &
Systems Div.
So. Norwalk, Conn.
56137 SpaL .Ing Fibre Co., Inc.
56289 Sprague Electric Co.
59446 Telex Corp.
59730 Thomas & Betts Co.
60741 Triplett Electrical Inst. Co.
61775 Union SWitch and Signal. Div. of
Westinghouse Air Brake Co.
Tonawanda, N. Y.
North Adams, Mass.
Tulsa, Okla.
Elizabeth, N. J.
Bluffton, Ohio
Pittsburgh, Pa.
62119 Universal Electric Co.
63743 Ward·Leonard Electric Co.
64959 Western Electric Co., Inc.
Owosso, Mich.
Mt. Vernon,. N. Y.
65092 Weston Inst. Inc. Weston-Newark
66295 Wittek Mfg. Co.
66346 Minnesota Mining & Mfg. Co.
New York, N. Y.
Newark, N. J.
Chicago, III.
Revere Mincom Div.
St. Paul, Minn.
70276 Allen Mfg. Co.
70309 Allied Control
70318 Allmelal Screw Product Co., Inc_
Hartford, Conn.
New York, N. Y.
70417 Amplex, Div. of Chrysler Corp.
70485 Atlantic India Rubber Works, Inc.
Garden City, N. Y.
Detroit, Mich.
Chicago, III.
70563 Amperrte Co., Inc.
70674 ADC Products Inc.
70903 Belden Mfg. Co.
Union City, N.
J.
Minneapolis, Minn.
Chicago, III.
Code
No.
Manufactuntr Address
70998 Bird Electronic Corp.
71002 Birnbach Radio Co.
71034 Bliley Electric Co., Inc.
Cleveland, Ohio
New York, N. Y.
Erie, Pa.
71041 Boston Gear Works Div. of Murray Co, of Texas Quincy, Mass.
71218 Bud Radio, Inc.
71279 Cambridge Thermionics Corp.
71286 Camloc Fastener Corp.
71313 Cardwell Condenser Corp.
Willoughby, Ohio
Cambridge, Mass,
Paramus, N.J.
Lindenhurst L.I., N. Y.
71400 Bussmann Mfg. Div. of McGraw-Edison Co.
St. Louis, Mo.
'1436 Chicago Condenser Corp.
71447 Calif. Spring Co., Inc.
71450 CTS Corp.
71468 ITT Cannon Electric Inc.
Chicago, III.
Pico-Rivera, Calif.
Elkhart, Ind.
71471 Cinema, Div. Aerovox Corp.
71482 C. P. Clare & Co.
Los Angeles, Calif
Burbank, Calif.
Chicago, III.
71590 Centralab Div. of Globe Union Inc.
Milwaukee, Wis_
71616 Commercial Plastics Co.
71700 Cornish Wire Co., The
Chicago, III.
New York, N. Y.
71707 Coto Coil Co., Inc.
71744 Chicago Miniature Lamp Works
Providence, R.1.
71785 Cinch Mfg. Co., Howard B. Jones Div.
Chicago, III.
71984 Dow Corning Corp.
Chicago, III.
Midland, Mich.
72136 Electro Motive Mfg. Co., Inc.
Willimantic, Conn.
72619 Dialight Corp.
Brooklyn, N. Y.
72656 Indiana General Corp., Electronics Div.
Keasby, N. J.
72699 General Instrument Corp., Cap. Div. Newark, N.
J.
72765 Drake Mfg. Co.
72825 Hugh H. Eby Inc.
Harwood Heights, III.
Philadelphia, Pa.
72928 Gudeman Co.
72962 Elastic Stop Nut Corp.
72964 Robert M. Hadley Co.
Chicago, III.
Union, N. J.
Los Angeles, Calif.
72982 Erie Technological Products, Inc.
Erie, Pa.
73061 Hansen Mfg. Co., Inc.
73076 H. M. Harper Co.
Princeton, Ind.
Chicago, III.
73138 Helipot D,v. of Beckman Inst., Inc.
Fullerton, Calif.
73293 Hughes Products DiVision of Hughes
Aircraft Co.
Newport Beach, Calif.
73445 Amperex Elect Co.
73506 Bradley Semiconductor Corp.
Hicksville, L.I., N. Y.
New Haven, Conn.
73559 Carling ElectriC, Inc.
73586 Circle F Mfg. Co.
Hartford, Conn.
Trenton, N. J.
73682 George K. Garrett Co., Div. MSL
Industries Inc.
Philadelphia, Pa.
73734 Federal Screw Products Inc.
73743 Fischer SpeCial Mfg. Co.
73793 General Industries Co., The
73846 Goshen Stamping & Tool Co.
73899 JFD Electronics Corp.
73905 Jennings Radio Mfg. Corp.
Chicago, III.
Cincinnali, Ohio
Elyria, Ohio
Goshen, Ind.
Brooklyn, N. Y.
San Jose, Calif.
73957 Groov·Pin Corp.
74276 Signalile Inc.
Ridgefield, N. J.
Neptune, N. J.
74455 J. H. Winns, and Sons Winchester, Mass.
74861 Industrial Condenser Corp.
Chicago, III.
74868 R. F. Products Division of Amphenol·Borg
Electronics Corp.
74970 E.
F. Johnson Co.
75042 International Resistance Co.
75263 Keystone Carbon Co., Inc.
75378 CTS Knights Inc.
75382 Kulka Electric Corporalion
75818 Lenz Electric Mfg. Co.
75915 Littlefuse, Inc.
76005 Lord Mfg. Co.
Danbury, Conn.
Waseca, Minn.
Philadelphia, Pa.
St. Marys, Pa.
Sandwich, III.
Mt. Vernon, N. Y.
Chicago, III.
Des Plaines, III.
Erie, Pa.
Cod.
No.
Manufacturer
76210 C. W. Marwedel San Francisco, Calif.
76433 General Instrument Corp., Micamold Division
76487 James Millen Mfg. Co., Inc.
76493 J. W. Miller Co.
Newark, N.
J.
Malden, Mass.
Los Angeles, Calif.
76530 Cinch-Monadnock, Div. of United Carr
Fastener Corp.
San Leandro, Calif.
76545 Mueller Eleclric Co.
76703 National Union
Cleveland, Ohio
Newark, N. J.
76854 Oak Manufacturing Co.
Crystal Lake, III.
77068 The Bendix Corp., Electrodynamics Div.
N. Hollywood, Calif.
77075 Pacific Metals Co.
San Francisco, Calif.
77221 Phanostran Instrument and Electronic Co.
South Pasadena, Calif.
77252 Philadelphia Steel and Wire Corp.
Philadelphia, Pa.
77342 American Machine & Foundry Co. Potter
& Brumfield Div.
Princeton, Ind.
78277
78283
78290
78452
78471
78488
78493
78553
78790
78947
79136
79142
79251
79727
77630 TRW Electronic Components Div.
Camden, N. J.
77638 General Instrument Corp., Reclifier Div.
Brooklyn, N. Y.
77764 Resistance Products Co_
77969 Rubbercraft Corp. of Calif.
Harrisburg, Pa.
Torrance, Calif.
78189 Shakeproof Division of Illinois Tool Works
Elgin, III.
Sigma
Signal Indicator Corp.
Struthers·Dunn Inc.
Thompson·Bremer
Ti Itey Mfg. Co.
& Co.
Stackpole Carbon Co.
Siandard Thomson Corp.
So. Braintree, Mass.
NewYork,N.Y.
Pitman, N. J.
Chicago, III.
San Francisco, Calif.
St. Marys, Pa.
Waltham, Mass.
Cleveland, Ohio Tinnerman Products, Inc.
Transformer Engineers
Ucinite Co.
Waldes Kohinoor Inc.
Veeder Root, Inc.
Wenco Mfg. Co.
San Gabriel, Calif.
Newtonville, Mass.
Long Island City, N. Y.
Hartford, Conn.
Chicago, III.
Conlinental·Will Electronics Corp.
Philadetphia, Pa.
79963 Zierick Mfg. Corp.
New Rochelle, N. Y.
80031 Mepco Division of Sessions Clock Co.
Morristown, N. J.
80120 Schnitzer Alloy Products Co.
Elizabeth, N. J.
80131 Electronic Industries Association.
Any brand
Tube meeting EIA Standards-Washington, DC.
80207 Unimax SWitch, Div. Maxon Electronics Corp.
80223 United Transformer Corp.
80248 Oxford Electric Corp.
Wallingford, Conn.
New York, N. Y.
Chicago, III.
80294 Bourns Inc.
Riverside, Calif.
80411 Acro Div. of Robertshaw Controls Co.
Columbus, Ohio
80486 All Star Products Inc.
80509 Avery Label Co.
80583 Hammarlund Co., Inc.
80640 Stevens, Arnold, Co., Inc.
80813 Dimco Gray Co.
81030 Internationat Instruments Inc.
81073 Grayhill Co.
81095 Triad Transformer Corp.
Defiance, Ohio
Monrovia, Calif.
New York, N. Y.
Boston, Mass.
Dayton, Ohio
Orange, Conn.
LaGrange, It I.
Venice, Calif.
81312 Winchester Elec. Div. Litton Ind., Inc.
Oakville, Conn.
81349 Military Specification
81483 International Rectifier Corp.
EI Segundo, Calif.
81541 Airpax Electronics, Inc.
Cambridge, Maryland
81860 Barry Controls, D,v. Barry Wright Corp.
Watertown, Mass.
82042 Carter Precision Electric Co.
Skokie, III.
00015-45
Revised: Aplll, 1968
From: FSC.
H4-1
H4-2
Handbook Supplements
Dated AUGUST
Dated NOV.
1966
1962
Appendix A Model403B
CODE LIST OF MANUFACTURERS (Cont'd)
Code
No.
MCII'Iufocturer Address
82047
82142
Sperti Faraday Inc., Copper Hewill
Electric Div.
Jeffers Electronics Division of Speer
Hoboken, N. J.
Carbon Co.
Du Bois, Pa.
Fairchild Camera & Inst. Corp. Space & Defense 82170
82209
82219
82376
82389
82647
82768
82866
81877
82893
83058
83086
System Div.
Maguire Industries, Inc.
Paramus, N. J.
Greenwich, Conn.
Sylvania Electric Prod. Inc.
Electronic Tube Division
Astron Corp.
Emporium, Pa.
East Newark, Harrison, N, J,
Switchcraft, Inc.
Chicago, III.
Metals & Controls Inc. Spencer Products
Phillips·Advance Control Co.
Research Products Corp.
Rotron Mfg. Co., Inc.
Vector Electronic Co.
Carr Fastener Co.
Allleboro, Mass.
Joliet, III.
Madison, Wis.
Woodstock, N. Y.
Glendale, Calif.
Cambridge, Mass.
83125
New Hampshire Ball Bearing, Inc.
Peterborough, N. H.
General Instrument Corp., Capacitor Div.
Darlington, S. C.
83148
83186
83298
83315
83324
83330
83332
83385
83501
ITT Wire and Cable Div.
Victory Eng. Corp.
Bendix Corp., Red Bank Div.
Hubbell Corp.
Rosan Inc.
Smith, Herman H., Inc.
Tech Labs
Central Screw Co.
Los Angeles, Calif.
Springfield, N. J.
Red Bank, N. J.
Mundelein, III.
Newport Beach, Calif.
Brooklyn, N. Y.
Palisade's Park, N. J.
Chicago, III.
Gavill Wire and Cable Co.
Div. of Amerace Corp.
Brookfield, Mass.
Burroughs Corp. Electronic Tube Div.
83594
83740
Plainfield, N. J.
Union Carbide Corp. Consumer Prod. Div.
83777
8382 I
83942
84171
84396
84411
84970
85454
85471
85474
85660
85911
86197
Model Eng. and Mfg., Inc.
Loyd Scruggs Co.
Aeronauticallnsl. & Radio Co.
Arco Electronics Inc.
A. J. Glesener Co., Inc.
TRW Capacitor Div.
Sarkes Tarzlan, Inc.
Boonton Molding Company
A. B. Boyd Co.
R. M. Bracamonte
Kolierl Kords, Inc.
Sea,...
~ss
&
Rubber Co.
Co.
New York, N. Y.
Huntington, Ind.
Clifton Precision Products Co., Inc.
Festus, Mo.
Lodi, N.J.
Great Neck, N. Y.
San Francisco, Calif.
Ogallala, Neb.
Bloomington, Ind.
Boonton, N.1.
San,lFrancisco, Calif.
Sall. Francisco, Calif.
Hamden, Gonn.
Chicago, III.
Clifton Heights, Pa.
86579 Precision Rubber Products Corp.
86684 RadiO Corp. of America, Electronic
Compo & Devices Div.
Dayton, Ohio
Harrison, N.J.
87034 Marco Industries Anaheim, Calif.
87216 Phil co Corporation (Lansdale Division)
Lansdale, Pa.
87473 Western Fibrous Glass Products Co.
San Francisco, Calif.
87664 Van Walers & Rogers Inc.
87930 Tower Mfg. Corp.
88140 Cutler-Hammer, Inc.
San Francisco, Calif.
Providence, R. I.
Lincoln, III.
Code
No.
MCII'Iufacturer Address
88220 Gould·National Balleries, Inc.
88698 General Mills, Inc.
89231 Graybar Electric Co.
89473 G. E. Distributing Corp.
SI. Paul, Minn.
Buffalo, N. Y.
Oakland, Calif.
Schenectady, N. Y.
89665 United Transformer Co.
90030 United Shoe Machinery Corp.
Chicago, III.
Beverly, Mass.
90179 US Rubber Co., Consumer Ind. & Plastics
Prod. Div.
Passaic, N. J.
90970 Bearing Engineering Co.
San Francisco, Calif.
91146 ITT Cannon Elect, Inc., Salem Div. Salem, Mass.
91260 Connor Spring Mfg. Co.
91345 Miller Dial & Nameplate Co.
San Francisco, Calif.
EI Monte, Calif.
Chicago, III.
91418 RadiO Materials Co.
91506 Augatlnc.
91637 Dale Electronics, Inc.
91662 Elco Corp.
91737 Gremar Mfg. Co., Inc.
Attleboro, Mass.
Columbus, Nebr.
Willow Grove, Pa.
Wakefield, Mass.
91827 K F Development Co.
91886 Malco Mfg. Co., Inc.
Redwood City, Calif.
91929 Honeywell Inc., Micro Switch Div.
Chicago, III.
Freeport, III.
95236
95238
95263
95265
95275
95348
95354
95566
9m2
95984
95987
91961 Nahm·Bros. Spllng Co.
92180 Tru·Con nector Corp.
Oakland, Calif.
Peabody, Mass.
92367 Elgeet Opltcal Co. Inc.
92607 Tensolite Insulated Wire Co., Inc.
Rochester, N. Y.
Tarrytown, N. Y.
92702
92966
93332
IMC Magnetics Corp.
Hudson Lamp Co.
Wesbury Long Island, N. Y.
Kearney, N. J.
Sylvania Electric Prod. Inc.
93369
93410
93632
93929
94137
94144
Semiconductor Div.
Robbins & Myers Inc.
Stevens Mfg. Co., Inc.
Waters Mfg. Co.
G.V. Controls
General Cable Corp.
Woburn, Mass.
Palisades Park, N. J.
Mansfield, Ohio
Culver City, Calif.
Livingston, N. J.
Bayonne, N. J.
94148
Raytheon Co., Compo Div., Ind.
Compo Operations
Scientific Electronics Producls, Inc.
Quincy, Mass.
94154
94197
Loveland, Colo.
Wagner Elect. Corp., Tung·Sol Div. Newark, N. J.
Curtiss·Wright Corp. Electronics Div.
East Paterson, N. J.
94222
94330
94375
94682
South Chester Corp.
Wire Cloth Products, Inc.
Automatic Metal Producls Co.
Worcester Pressed Aluminum Corp.
Chester, Pa.
Bellwood, III.
Brooklyn, N. Y.
94696
95023
Worcester, Mass.
Magnecraft Electric Co.
George A. Philbrick Researchers, Inc.
Chicago, III.
Allies Products Corp.,
Continental Connector Corp.
Leecraft Mfg. Co., Inc.
Nalional Coil Co.
Vitramon, Inc.
Gordos Corp.
Methode Mfg. Co.
Arnold Engineering Co.
Dage Electric Co., Inc.
Siemon Mfg. Co.
Weckesser Co.
Boston, Mass.
Dania, Fla.
Woodside, N. Y.
Long Island, N. Y.
Sheridan, Wyo.
Bridgeport, Conn.
Bloomfield, N.J.
Rolling Meadows, III.
Marengo, III.
Franklin, Ind.
Wayne, III.
Chicago, III.
Code
No.
MCII'Iufoctur., Aclclress
96067 Huggins Laboratories
96095 Hi·Q Div. of Aerovox Corp.
96256 Thordarson·Meissner Inc.
96296 Solar Manufacturing Co.
96330 Carllon Screw Co.
96341 Microwave Associates, Inc.
96501 Excel Transformer Co.
97464 Industrial Retaining Ring Co.
97539 Automatic & Precision Mfg.
97979 Reon Resistor Corp.
Sunnyvale, Calif.
Olean, N. Y.
Mt. Carmel, III.
Los Angeles, Calif.
Chicago, III.
Burlington, Mass.
Oakland, Calif.
Irvington, N. J.
Englewood, N.J.
Yonkers, N.Y.
97983 Lillon System Inc., Adler·Westrex
Commun. D,v.
New Rochelle, N. Y.
98141 R·Troncis, Inc.
98159 Rubber Teck, Inc.
Jamaica, N. Y.
Gardena, Calif.
98220 Hewlett·Packard Co. , Moseley Div.
Pasadena, Calif.
98278 Microdot, Inc.
98291 Sealectro Corp.
98376 Zero Mfg. Co.
98731 General Mills Inc.,
So. Pasadena, Calif.
Mamaroneck, N. Y.
Electronics Div.
Burbank, Calif.
Minneapolis, Minn.
98734 Paeco Div. of Hewlell·Packard Co.
Palo Alto, Calif.
98821 North Hills Electronics, Inc.
Glen Cove, N. Y.
98978 International Electronic Research Corp.
Burbank. Calif.
99109 Columbia Technical Corp.
99313 Varian Associates
99378 Attee Corp.
99515 Marshall Ind., Capacitor Div.
New York, N. Y.
Palo Alto, Calif.
Winchester, Mass.
Monrovia, Calif.
99707 Control Switch Division, Controls Co.
of America EI Segundo, Calif.
99800 Delevan Electronics Corp.
99848 Wilco Corporation
99934 Renbrandt, Inc.
99942 Hoffman Electronics Corp.
Semiconductor Div.
East Aurora, N. Y.
Indianapolis, Ind.
Boston, Mass.
EI Monte, Calif.
99957 Technology Instrumenl Corp. of Calif.
Newbury Park, Calif.
THE FOLLOWING HP VENDORS HAVE NO NUMBER
ASSIGNED IN THE LATEST SUPPLEMENT TO THE
FEDERAL SUPPLY CODE FOR MANUFACTURERS
HANDBOOK.
OOOOF
OOOOZ
Malco Tool and Die Los Angeles, Calif.
Willow Leather Products Corp.
Newark, N. J.
OOOAB
OOOBB
ETA
PreCISion Instrument Components Co.
England
Van Nuys. Calif.
OOOCS Hewlett-Packard Co., Colorado Springs.
OOOMM
OOONN
OOOQQ
OOOWW
Cotorado Springs, Colorado
Rubber Eng. & Development Hayward, Calif.
A "N" 0 Mfg_ Co_
Coollron
San Jose, Calif.
Oakland, Calif.
California Easlern Lab.
ooon
S. K. Smllh Co.
Burlinglon, Calif.
Los Angeles, Calif.
SUPPLEMENTAL CODE LIST OF MANUFACTURERS
OOOLM Newark
13189 Bunt. Metal Product.
Co.
'4163 Phelpe Dodce Corp.
08300 $one Paper Tube Co.
OUII Skottte
Denver, Colorado
Chicaeo.
mlnal.
New York. New York
WUhlngton. D. C.
Archibald.
Penn.ylvanLJ
00015-45
Revised: Aplll, 1968
From: FSC.
H4-1
H4-2
Handbook Supplements
Dated AUGUST 1966
Dated NOV 1962
Mode1403B Appendix B
SALES & SERVICE OFFICES
UNITED STATES
ALABAMA
P.O. Box 4207
2003 Byrd Sprine Road S.W.
Hlnlsville 35802
Tel: (205) 881·4591
TWX: 810·726·2204
ARIZONA
2336 E.
Magnolia SI.
Phoenix 85034
Tel: (602) 252·5061
TWX: 910·951·1330
5737 East Broadway
Tucson 85716
Tel: (602) 298·2313
TWX: 910·952·1162
CONNECTICUT
508 Tolland Street
East Hartford 06108
Tel: (203) 289·9394
TWX: 710·425·3416
111 East Avenue
Norwalk 06851
Tel: (203) 853·1251
TWX: 710·468-3750
FLORIDA
P.O. Box 24210
2806 W. Oakland Park Blvd.
FI. lauderdale 33307
Tel: (305) 731·2020
TWX: 510·955·4099
P.O. Box 20007
Herndon Station 32814
621 Commonweallh Avenue
Orlando
Tel: (305) 841-3970
TWX: 810-850'0113
CALIFORNIA
1430 East Orangethorpe Ave.
Fullerton 92631
Tel: (714) 870·1000
3939 Lankerehim Boulevard
North Hollywood 91604
Tel: (213) 877·1282
TWX: 910·499·2170
1101 Embarcadero Road
Palo Allo 94303
Tel: (415) 327·6500
TWX: 910·373·1280
2220 Wall Ave.
Sacramento 95825
Tel: (916) 482·1463
TWX: 910·367·2092
9606 Aero Drive san 01'10 92123
Tel: (714) 279·3200
TWX: 910·335·2000
COLORAOO
7965 East Prentice
£nllewood 80110
Tel: (303) 771-3455
TWX: 910-935'0705
GEORGIA
P.O. Box 28234
450 Interstate North
Allanta 30328
Tel: (404) 436·6181
TWX: 810-766-4890
ILLINOIS
5500 Howard Street
Skokie 60076
Tel: (312) 677-0400
TWX: 910-223·3613
INDIANA
3839 Meadows Drive
Indianapolis 46205
Tel: (317) 546-4891
TWX: 810-341-3263
LOUISIANA
P.O_ Box 856
1942 Williams Boulevard
Ilenner 70062
Tel: (504) 721·6201
TWX: 810-955-5524
MARYLAND
6707 Whitestone Road
Baltimore 21207
Tel: (301) 944·5400
TWX: 710-862-9157
P.O. Box 1648
2 Choke Cherry Road
Rockville 20850
Tel: (301) 948·6370
TWX: 710-828-9684
MASSACHUSETTS
32 Hartwell Ave.
Lexinlton 02173
Tel: (617) 861-8960
TWX: 710-326·6904
MICHIGAN
21840 West Nine Mile Road
Southfield 48075
Tel: (313) 353-9100
TWX: 810·224·4882
MINNESOTA
2459 University Avenue
St_ Paul 55114
Tel: (612) 645-9461
TWX: 910·563·3734
MISSOURI
11131 Colorado Ave.
Ilansas City 64137
Tel: (816) 763-8000
TWX: 910-771-2087
2812 South Brentwood Blvd.
SI. Louis 63144
Tel: (314) 962-5000
TWX: 910·760-1670
NEW JERSEY
W. 120 Century Road
Paramus 07652
Tel: (201) 265-5000
TWX: 710·990·4951
1060 N. Kings Highway
Cherry Hill 08034
Tel: (609) 667-4000
TWX: 710-892-4945
NEW MEXICO
P.O. Box 8366
Station C
6501 Lomas Boulevard
Albuquerque 87108
Tel: (505) 265-3713
TWX: 910-989-1665
H.E.
156 Wyall Drive
Las Cruces 88001
Tel: (505) 526-2485
TWX: 910-983-0550
NEW YORK
1702 Central Avenue
Albany 12205
Tel: (518) 869-8462
TWX: 710-441·8270
1219 Campville Road
Endicott 13760
Tel: (607) 754-0050
TWX: 510-252-0890
82 Washington Street
Poulhkeepsle 12601
Tel: (914) 454-7330
TWX: 510·248·0012
39 Saginaw Drive
Rochester 14623
Tel: (716) 473-9500
TWX: 510-253-5981
5858 East Molloy Road
Syracuse 13211
Tel: (315) 454-2486
TWX: 710-541-0482
1 Crossways Park West
Woodbury 11797
Tel: (516) 921-0300'
TWX: 510-223-0811
NORTH CAROLINA
P.O. Box 5188
1923 North Main Street
Hilh Point 27262
Tel: (919) 885-8101
TWX: 510·926-1516
OHIO
25575 Center Ridee Road
Cleveland 44145
Tel: (2)6) 835-0300
TWX: 810-427-9129
3460 South Dixie Drive
Dayton 45439
Tel: (513) 298-0351
TWX: 810·459-1925
1120 Morse Road
Columbus 43229
Tel: (614) 1146·1300
OKLAHOMA
2919 United Founders Boulevard
Oklahoma City 73112
Tel: (405) 848-2801
TWX: 910-830·6862
OREGON
Westhilis Mall, Suite 158
4475 S.W. Scholls Ferry Road
Porlland 97225
Tel: (503) 292-9171
TWX: 910-464·6103
PENNSYLVANIA
2500 Moss Side Boulevard
Monroeville 15146
Tel: (412) 271·0724
TWX: 710-797-3650
1021 8th Avenue
Kine of Prussia Industrial Park
Ilinl of Prussia 19406
Tel: (215) 265-7000
TWX: 510-660-2670
VERMONT
P.O. Box 2287
Kennedy Drive
South Burllneton 05401
Tel: (802) 658-4455
TWX: 510·299·0025
VIRGINIA
P.O. Box 6514
2111 Spencer Road
Richmond 23230
Tel: (703) 285-3431
TWX: 710-956-0157
WASHINGTON
433·108th N.E.
Bellevue 98004
Tel: (206) 454·3971
TWX: 910-443-2303
·WEST VIRGINIA
Charleston
Tel: (304) 768·1232 RHODE ISLAND
873 Waterman Ave.
East Providence 02914
Tel: (401) 434-5535
TWX: 710-381-7573
P.O_ Box 22813
6300 Westpark Drive
Suite 100
Houston 77027
Tel: (713) 781-6000
TWX: 910-881-2645
231 Billy Mitchell Road
San Antonio 78226
Tel: (512) 434-4171
TWX: 910·871-1170
UTAH
2890 South Main Street
Salt Lake City 84115
Tel: (80l) 487·0715
TWX: 910-925-5681
TEXAS
P.O. Box 1270
201 E. Arapaho Rd.
Richardson 75080
Tel: (214) 231-6101
TWX: 910-867-4723
FOR U.S. AREAS NOT
LISTED:
Contact the regional office nearest you: Allanta, Georgia
North Hollywood, California
Paramus, New Jersey ... Skokie,
Illinois.
Their Jmplete addresses are Iisteo above.
.
.
·Service Only
CANADA
ALBERTA
Hewlell-Packard (Canada) Ltd.
11745 Jasper Ave.
Edmonton
Tel: (403) 482-5561
TWX: 610-831-2431
BRITISH COLUMBIA
Hewlell-Packard (Canada) Ltd.
4519 Canada Way
North Burnaby 2
Tel: (604) 433-8213
TWX: 610-922·5059
MANITOBA
Hewlell-Packard (Canada) L d.
511 Bradford CI.
Winnipeg
Tel: (204) 786·7581
TWX: 610·671-3531
NOVA SCOTIA
Hewlell-Packard (Canada) Ltd.
2745 Dutch Village Rd.
Suite 206
Halifax
Tel: (902) 455-0511
TWX: 610-271·4482
ONTARIO
Hewlell·Packard (Canada) Ltd.
880 Lady Ellen Place
Ollawa 3
Tel: (613) 722-4223
TV/X 610·562-1952
HeNlell-Packard (Canada) Ltd.
50 Galaxy Blvd.
Rexdale
Tel: (416) 677·9611
TWX: 610·492-4246
QUEBEC
Hewlell-Packald (Canada) Ltd
275 Hymus Boulevard
Pointe Claire
Tel (514) 697·4232
TWX· 610·422-3022
Telex: 01·20607
FOR CANADIAN AREAS NOT
LISTED:
Contact Hewlell·Packard (Canada) Ltd. in Pointe Claire, at the complete address listed above.
CENTRAL AND SOUTH AMERICA
ARGENTINA
Hewlett-Packard Argentina
S.A.C.e.1
Lavalle 1171 - 3
8uenos Aires
Tel: 35·0436, 35·0627, 35·0431
Telex: 012·1009
Cable. HEWPACKARG
Hewlell·Packard Do Brasil
Industria e Comercia llda.
Praca Dam Feliciano 78
Salas 806-8
Porto Alegre RGS
ECUADOR
Laboratorios de Radio-Ingenieria
Calle GuayaQuil 1246
Post Ollice Box 3199 auito
Tel: 212·496; 219·185
Cable: HORVATH Quito
BRAZIL
Hewlell·Packard Do Brasil l.e.C Ltda.
Rua Frei Caneca 1119
Sao Paulo· 3, SP
Tel: 288·7111, 287·5858
Cable: HEWPACK Sao PaulO
CHILE
H~ctor Calcagni y Cia, llda.
Bustos, 1932-3er Piso
Casilla 13942
Sanliaeo
Tel: 423 96
Cable: CALCAGNI Santiago
EL SALVADOR
Electr6nica
Apartado Postal 1589
Blvd. Venezuela 1231
San Salvador
Tel: 217527; 214895
Cable: ELECTRONICA
San Sal ador
Hewlell-Packard Do Brasil
Praca Dam Feliciano 78
Salas 806/808
Porto Alegre
Rio Crande do Sui (RS)-Brasil
Tel: 25·8470
Cable HEWPACK Porto Alegre
COLOMBIA
Instrumentacion
Henrik A. Langebaek llda.
Carrera 7 No. 48·59
Apartado Aereo 6287
& K,er
BOlota, 1 D.£.
Tel: 45-78-06, 45·55-46
Cable: AARIS Bogota
Telex: 44400 INSlCO
Hewlell·Packard Do BraSIl l.e.C. Ltda.
Rua da Matriz 29
80tafogo ZC·02
Rio de Janeiro, GB
Tel: 246-4417
Cable: HEWPACK RIO de laneoro
COSTA RICA
Lic. Allredo Gallegos Gurdi~n
Apartado 10159
San Jose
Tel: 21-86-13
Cable: GALGUR San Jos~
MEXICO
Hewlet ·Packard Mexicana, SA.
de C.V.
622 Adalia Prieto
Col. del Valle
Mexico 12. D.F.
Tel: 541·4232; 523-1874
Telex: 0017·74507
NICARAGUA
Roberto Ter~n G.
Apartado Postal 689
Edificio Ter~n
Managua
Tel. 3451. 3452
Cable: ROTE RAN Managua
PANAMA
Electr6nico 8alboa, S.A.
P.O. Box 4929
Ave. Manuel Espinosa No. 13-50
Bldg. Alina
Panama City
Tel: 230833
Telex' 3481003, Curundu.
Canal Zone
Cable: ELECTRON Panama City
PERU
Campania Electro MedIca S.A.
Ave. EnriQue Canaual 312
San Isidro
Casilla 1030
Lima
Tel: 22·3900
Cable: ELMED Lima
PUERTO RICO
San Juan Electronics, Inc.
P.O. Box 5167
Ponce de Leon 154
Pda. 3·PTA de Tierra
San Juan 00906
Tel: (809) 725·3342, 722·3342
Cable: SATRONICS San Juan
Telex· SATRON 3450 332
SURINAME
Sur tel-Radio Holland N.V.
P.O. Box 155
Paramaribo
Tel: 72118
Cable: T,eurnlet Paramaribo
URUGUAY
Pablo Ferrando S.A.
Comercial e Industrial
A enid a Italia 2877
Casilia de Correa 370
Montevideo
Tel: 40·3102
Cable: RADIUM Montevideo
VENEZUELA
Hewlett·Packard De Venezuela
C.A.
Apartado 50933
Caracas
Tel: 71.88.05, 71.88.69, 71.99.30
Cable: HEWPACK Caracas
Telex: 39521146
FOR AREAS NOT LISTED,
CONTACT:
Hewlett·Packard
INTERCONTINENTAL
3200 Hillview Ave.
Palo Alto, California 94304
Tel: (415) 493-1501
TWX: 910-373-1267
Cable: HEWPACK Palo Alto
Tele.: 034-8461
E 6·71
Appendix B Mode1403B
EUROPE
AUSTRIA
Unitabor GmbH
Wissenschaftliche Instrumente
Rummelhardtcasse 6
P.O. Box 33
A·1095 Vienna
Tel: (222) 42 6181,431394
Cable: LABOR INSTRUMENT
Vienna
Telex: 75 762
BELGIUM
Hewlett·Packard S.A. Benelux
348 Boulevard du Souverain
B·1I60 Brussels
Tel: (02) 722240
Cable: PA10BEN Brussels
Telex: 23 494
DENMARK
Hewlell·Packard A/S
Oatavej 38
OK·3460 Birkerod
Tel: (01) 81 66 40
Cable: HEWPACK AS
Telex: 66 40
Hewlell·Packard A/S
Torvet 9
OK·8600 Silkeborc
Te': (06)·82·71·66
FINLAND
Hewlell·Packard Oy
Bulevardi 25
P.O. Box 12185
Helsinki 12
Tel: 13·730
Cable: HEWPACKOY·Helsinkl
Telex: 12·1563
FRANCE
Hewlell·Packard France
Quartier de Courtaboeuf
Bolle Postale No.6
F·91 Orsay
Tel: 1-920 B8 01
Cable: HEWPACK Orsay
Telex: 60048
Hewlett·Packard France
4 Qual des Etroits
F·69 lyon 5~me
Tel: 78·42 63 45
Cable: HEWPACK lyon
Telex: 31617
Hewlett·Packard France
29 rue de la Gara
F·31 Blacnac
Tel: (61) 85 82 29
Telex: 51957
GERMANY
Hewlett·Packard Verlriebs·GmbH
Berliner Strasse 117
Poslfach 560/40
06 Nieder·Eschbach/Ffm 56
Tel: (0611) 50 10 64
Cable: HEWPACKSA Frankfurt
Telex: 41 32 49 FRA
Hewlett·Packard Vertriebs·GmbH
Wilmersdorfer Strasse 113/114
0·1000 Berlin W. 12
Tel: (0311) 3137046
Telex: 18 34 05
Hewlett·Packard Vertriebs·GmbH
Herrenbercerstrasse 110
07030 Biiblincen, Wiirllemberg
Tel: (07031) 66 72 86
Cable: HEPAG Bliblingen
Telex: 72 65 739
Hewlett·Packard Vertriebs·GmbH
Vogelsancer Weg 38
04 Dusseldorf
Tel: (0211 63 80 31/35
Telex: 85/86 533
Hewlett·Packard Vertrlebs·GmbH
Wendenstr. 23
02 Hamburc 1
Tel: (0411) 24 05 51/52
Cable: HEWPACKSA Hamburc
Telex: 21 5332
Hewlett·Packard Verlrlebs·GmbH
Reginfriedstrasse 13
08 Munchen 9
Tel: (0811) 69 59 71/75
Cable: HEWPACKSA Munchen
Telex: 52 49 8~
GREECE
Kostas Karayannis
18, Ermou Street
Athens 126 \
Tel: 230301,3,5
Cable: RAKAR Athens
Telex: 21 59 62 RKAR GR
IRELAND
Hewlett·Packard ltd.
224 Bath Road
SlOUCh, Sli 4 OS, Bucks
Tel: SlOUCh 753·33341
Cable: HEWPIE SlOUCh
Telex: 84413
ITALY
Hewlett·Packard Italiana S.p.A.
Via Amerigo Vespucci 2
1·20124 Mitan
:rei: (2) 6251 (10 lines)
Cable: HEWPACKIT Milan
Telex: 32046
Hewlett·Packard lIaliana S.p.A.
Palazzo lIalia
Piazza Marconi 25
1·00144 Rome· Eur
Tel: 6·591 2544
Cable: HEWPACKIT Rome
Telex: 61514
NETHERLANDS
Hewlett·Packard Benelux, N.V.
Weerdestein 117
P.O. Box 7825
Amsterdam, Z 11
Tel: 020·42 77 77
Cable: PA10BEN Amsterdam
Telex: 13 216
NORWAY
Hewlett·Packard Norce A/S
Box 149
Nesve,en 13
N·1344 Haslum
Tel: (02)·53 83 60
Cable: HEWPACK Oslo
Telex: 16621
PORTUGAL
Telectra
Empresa Tecnica de
EQuipamentos
Electricos, S.a.r.1.
Rua Rodrico da Fonseca 103
P.O. Box 2531 lisbon 1
Tel: 68 60 72
Cable: TE1ECTRA lisbon
•
Telex: 1598
SPAIN
Ataio Ingenieros SA
Enrique larreta 12
Madrid, 16
Tel: 215 35 43
Cable: TElEATAIO Madrid
Telex: 27249E
Ataio Ingenieros SA
Ganduxer 76
Barcelona 6
Tel: 211·44·66
Cable: TElEATAIO Barcelona
SWEDEN
Hewlett·Packard Sveriae AB
Enichetsv~cen 1·3
Fack
S·161 20 8romma 20
Tel: (08) 98 12 50
Cable: MEASUREMENTS
Stockholm
Telex: 10721
Hewlett·Packard Sverige AB
Hacakerscatan 9C
S·431 41 Millndal
Tel: 031 ·276800
Telex: 21 312 hpmlndl
SWITZERLAND
Hewlett Packard Schweiz AG
Zurcherstrasse 20
CH·8952 Schlieren Zurich
Tel: (051) 98 18 21/24
Cable: HPAG CH
Telex: 53933
Hewlett Packard Schweiz A.G.
Rue du Bois·du·lan 7
1217 Meyrin 2 Geneva
Tel: (022) 41 5400
Cable: HEWPACKSA Geneva
Telex: 27333 HPSA CH
TURKEY
Telekom Encineering Bureau
P.O. Box 376
Karakoy
Istanbul
Te': 49 40 40
Cable: TE1EMATlON Istanbul
UNITED KINGDOM
Hewlett·Packard ltd.
224 Bath Road
SlOUCh, Sli 4 OS, Bucks
Tel: Slough (0753) 33341
Cable: HEWPIE SlOUCh
Telex: 84413
Hewlett·Packard ltd.
The Graftons
Stamford New Road
Allrincham, Cheshire
Tel: 061 928·8626
Telex: 668068
YUGOSLAVIA
Belram SA
83 avenue des Mimosas
Brussels 1150, 8elcium
Tel: 34 33 32, 34 26 19
Cable: 8E1RAMEl Brussels
Telex: 21790
SOCIALIST COUNTRIES
PLEASE CONTACT:
Hewlett·Packard Ges.m.b.H
Innstrasse 23/2
Postfach
A1204 Vienna, Austria
Tel: (222) 3366 06/09
Cable: HEWPACK Vienna
Telex: 75923
ALL OTHER EUROPEAN
COUNTRIES CONTACT:
Hewlett·Packard S.A.
Rue du Bois·du·lan 7
1217 Meyrin 2 Geneva
Switzerland
Tel: (022) 41 54 00
Cable: HEWPACKSA Geneva
Telex: 2.24.86
AFRICA, ASIA, AUSTRALIA
ANGOLA
Telectra Empresa Hcnia de EQuipamentos EI~clricos
SAR
Rua de Barbosa Rodricues
42·1·
Box 6487 luanda
Cable: TElECTRA luanda
CEYLON
United Electrlcals ltd.
P.O. Box 681
Yahala Buitdlnc
Staples Street
Colombo 2
Tel: 5496
Cable: HOTPOINT Colombo
AUSTRALIA
Hewlett·Packard Australia
Pty. ltd.
22·26 Weir Street
Glen Iris, 3146
Victoria
Tel: 20.1371 (6 lines)
Cable: HEWPARO Melbourne
Telex: 31024
Hewlett·Packard Australia
Pty. ltd.
61 Alexander Street
Crows Nest 2065
New South Wales
Tel: 43.7866
Cable: HEWPARO Sydney
Telex: 21561
Hewlett·Packard Auslralia
Ply. ltd.
97 Churchill Road
Prospect 5082
South Australia
Tel: 65.2366
Cable: HEWPARO Adelaide
Hewlett Packard Australia
Pty. ltd.
2nd Floor, Suite 13
Casablanca Buildlncs
196 Adelaide Terrace
Perth, W.A. 6000
Tel: 21·3330
Cable: HEWPARO Perth
Hewlett·Packard Australia
Pty. ltd.
10 Woolley Street
P.O. Box 191
Dickson A.C.T. 2602
Tel: 49·8194
Cable: HEWPARO Canberra ACT
Hewlett·Packard Australia
Pty. lid.
75 Simpsons Road
Bardon
Queensland, 4068
Tel: 36·5411
Blue Star lid.
Blue Star House,
34 Ring Road lajpat Nacar
New Deihl 24, India
Tel: 62 32 76
Tel": 463
Cable: B1UESTAR
CYPRUS
Kypronlcs
19 Grecorios &
P.O. Box 1152.
Xenopoulos Road
Blue Star ltd.
17·C Ulsoar Road
Banaalore·8
Nicosia
Tel: 6282·75628
Cable: HE·I·NAMI
Blue Star. ltd.
7 Hare Street
P.O. Box 506
Calcutta 1, India
Tel: 23·0131
Telex: 655
Cable: BlUESTAR
Blue Star, ltd.
96 Park lane
Secunderabad 3, India
Tel: 7 63 91
Cable: B1UEFROST
ETHIOPIA
African Salespower & Agency
Private ltd., Co.
P. O. Box 718
58/59 Cunningham SI.
Addis Ababa
Blue Star, ltd.
23/24 Second line Beach
Madras I, India
Tel: 2 39 55
Telex: 379
Cable: B1UESTAR
Tel: 12285
Cable: ASACO Addlsababa Blue Star, lid.
& Co. (Honc Kong) ltd.
1B Kaiser Bungalow
HONG KONG
Oindli Road
Schmidt
Jamshedpur, India
P.O. Box 297
Tel: 38 04
1511, Prince's BuildinC 15th Floor Cable: B1UESTAR
10, Chater Road
Hone Kone
Tel: 240168, 232735
Cable: SCHMIDTCO Hong Kane
INDONESIA
Bah Bolon Tradlna Coy. N.V.
Olalah Merdeka 29
Bandune
INDIA
Blue Star ltd.
Kasturl BUildincs
Tel: 4915; 51560
Cable: 11MU
Telex: 08·809
Jamshedji Tata Rd.
Bombay 20BR, India
IRAN
Tel: 29 50 21
Telex: 2156
Cable: BlUEFROST
Telecom, ltd.
P. O. Box 1812
240 Kh. Saba Shomall
Blue Star ltd.
Teheran
Tel: 43850, 48111
Band Box House
Prabhadevi
Cable: BASCOM Teheran
Bombay 2500, India
Tel: 45 73 01
Telex: 2156
ISRAEL
Electronics & Encineerlng
Olv. of Motorola Israel ltd.
Cable: BlUESTAR
17 Amlnadav Street
Tel·Aviv
Blue Star ltd.
14/40 Civil lines
Tel: 36941 (3 lines)
Kanpur, India
Tel: 688 82
Cable: BASTEI Tel·Avlv
Telex: Bastel Tv 033·569
Cable: B1UESTAR JAPAN
Yokogawa·Hewlett·Packard ltd.
Ohashi Buildinc
1·59·1 Yoyogi
Shibuya·ku, Tokyo
Tel: 03·370·2281/7
Telex: 232·2024YHP
Cable: YHPMARKET TOK 23·724
Yokocawa·Hewlett·Packard ltd.
Nisei Ibaragi BldC.
2·2·8 Kasuca
Ibaraci·Shi
Ouka
Tel: (0726) 23·1641
Telex: 385·5332 YHPOSAKA
Yokocawa·Hewlett·Packard ltd.
Ito Bulldlnc
No. 59, Kotorl·cho
Nakamura·ku, Nacoya City
Tel: (052) 551·0215
Yokogawa·Hewlett·Packard ltd.
Nitta Bldg.
2300 Shinohara·cho,
Kohoku·ku
Yokohama 222
Tel: (405) 432·1504/5
KENYA
Kenya Kinetics
P.O. Box 18311
Nairobi, Kenya
Tel: 57726
Cable: PROTON
KOREA
American Tradinc Co.,
Korea, ltd.
Seoul P.O. Box 1103
7th & 8th floors, OaeKyuna Bldg.
107 Sejone Ro
Chongro·Ku, Seoul
Tel: 75·5841 (4 lines)
Cable: AMTRACO Seoul
LEBANON
Constantin E. Macridls
Clemenceau Street
P.O. Box 7213
Beirut
Tel: 220846
Cable: ELECTRO NUCLEAR Beirut
MALAYSIA
MECOMB Malaysia ltd.
2 loronc 13/6A
Section 13
Petallnc Jaya, Selaneor
Cable: MECOMB Kuala lumpur
MOZAMBIQUE
A. N. Goncalves, IDA.
4.1 ApI. 14 Av. O. luis
P.O. Box 107 lourenco Marques
Cable: NEGON
NEW ZEALAND
Hewlett·Packard (N.Z.) ltd.
94·96 Dixson SI.
P.O. Box 9443
Wellineton, N.Z.
Tel: 56·559
Cable: HEWPACK Wellincton
Hewlett Packard (N.Z.) ltd.
Box 51092
Pukuranca
Tel: 573·733
Cable: HEWPACK, Auckland
PAKISTAN (EASn
Mushko & Company, ltd.
I, Jinnah Avenue
Dacca 2
Tel: 280058
Cable: NEWOEAl Dacca
PAKISTAN (WEST)
Mushko & Company, ltd.
Oosman Chambers
Abdullah Haroon Road
KarachI 3
Tel: 511027, 512927
Cable: COOPERATOR Karachi
PHILIPPINES
Electromex Inc.
Makatl Commercial Center
2129 Pasonc Tamo
Makati, Rlzal 0 708
P.O. Box d028
Manila' -,.
Tel: 89·85·01; 88·91-71
Caf\e;-'ElEMEX Manila
SINGAPORE
Mechanical and Combustion
Enclneerina Company ltd.
9, Jalan Kllanc
Red Hill Industrial Estate
Sincapore, 3
Tel: 642361·3; 632611
Cable: MECOMB Sincapore
Hewlett·Packard Far East
Area Office
P.O. Box 87
Alexandra Post Office
Sineapore 3
Tel: 633022
Cable: HEWPACK SINGAPORE
SOUTH AFRICA
Hewlett Packard South Africa
(Pty.), ltd.
P.O, Box 31716
Braamfontein Transvaal
Milnerton
30 De Beer Street
Johannesburc
Tel: 725·2080, 725·2030
Telex: 0226 JH
Cable: HEWPACK Johannesbura
Hewlett Packard South Africa
(Pty.), ltd.
Breecastle House
Bree Street
Cape Town
Tel: 3·6019, 3·6545
Cable: HEWPACK Cape Town
Telex: 5·0006
Hewlett Packard South Africa
(Pty.), ltd.
30B Glenwood Centre
Corner Hunt
P.O. Box 99
& Moore Roads
Dverport, Natal
Tel: 347536
TAIWAN
Hewlett Packard Taiwan
39 Chung Shiao West Road
Sec. 1
Overseas Insurance
Corp. 81dC. 7th Floor
Taipei
Tel: 579·605, 579-610, 579·613
Telex: TP824 HEWPACK
Cable: HEWPACK Taipei
THAILAND
The International
Enclneerina Co., ltd.
P. O. Box 39
614 Sukhumvit Road
Banekok
Tel: 910722 (7 lines)
Cable: GYSOM
T1X INTENCO BK·226 Bangkok
UGANDA
Uganda Tele·Eleclric Co., ltd.
P.O. Box 4449
Kampala
Tel: 57279
Cable: COMCO Kampala
VIETNAM
Peninsular Trading Inc.
P.O. Box H·3
216 Hlen·Vuong
Saleon
Tel: 20805, 93398
Cable: PENTRA, SAIGON 242
ZAMBIA
R. J. Tilbury (Zambia) Ltd.
P.O. Box 2792 lusaka
Zambia, Central Africa
Tel: 73793
Cable: ARJAYTEE, lusaka
MEDITERRANEAN AND
MIDDLE EAST COUNTRIES
NOT SHOWN PLEASE
CONTACT:
Hewlett·Packard Correspondence
Office
Piazza Marconi 25
1·00144 Rome·Eur, Italy
Tel: (6) 59 40 29
Cable: HEWPACKIT Rome
Telex: 61514
OTHER AREAS NOT
LISTED, CONTACT:
Hewlett·Packard
INTERCONTINENTAL
3200 Hillview Ave.
Palo Alto, California 94304
Tel: (415) 326·7000
(Feb. 71 493-1501)
TWX: 910·373·1267
Cable: HEWPACK Palo Alto
Telex: 034·8461
Model403B
Appendix C
II
MANUAL BACKDATING CHANGES
MODEL 403B
TRANSISTORIZED AC VOLTMETER
This manual backdating sheet makes this manual appllcable to earlier instruments.
Instrument-component values that differ from those in the manual, yet are not listed in the backdating sheet, should be replaced using the part number given in the manual.
Instrument Serial Prefix Make Manual Changes Instrument Serial Prefix Make Manual Changes
225-
523-05300 and below
523-05825 and below
523-09985 and below
CHANGE 1:
CHANGE 2:
CHANGE 3:
CHANGE 4:
CHANGE 5:
CHANGE 6:
1
2
4
3
949-10635 and below 5
Table 6-1, Page 6-6:
Delete index #19, Battery Holder; -hp- Part No. 5040-0615; 2 ea.
Add Battery Holder; -hp- Part No. 403B-64A and 403B-64B.
Page 6-2:
Change Q7 from 2N3638 (1853-0016) to 2N1183 (1850-0064).
- - - - - - - N O T E - - - - - - - - - -
Transistor 2N3638 is a preferred change for all
403B instruments.
Page 5-11/5-12, Figure 5-8:
Change C15 from: 100 pF, Part No. 0146-0176 to: 120 pF, Part No. 0140-0216
Table 6-1:
Change the following Part Nos:.
DS1 to 1450-0048, 32 to 1251-0148, W1 to 8120-0078.
Table 6-2:
Change the following Part Nos:
Item 9 to 403B-2B, Item 23 (Fig. 6-1) to 403B-2B.
Pages 5-9, 6-2, 6-3:
Change the following parts descriptions:
Q1 to 5080-9029; TSTR Ge PNP Selected
Q2, 3. and 5 to 1850-0096; TSTR Ge PNP 2N2189
Q4, 6 to 1854-0017; TSTR Si NPN 2N706A
R7 to 0758-0051; R: fxd comp 43K ± 5% 1/2 W
The voltages printed in red do not apply to these instruments.
The covers, panels, and trim listed in the parts list are for brown instruments.
Covers, panels. and trim for blue instruments are listed below:
403B-2A Front Panel
5060 -0705
5000-0702
5000-5388
5020-5388
Top Cover Assy
Side Cover
Bottom Cover
Meter Trim
·:MANUAl CHANGES
Model 4038
AC VOLTMETER
Manual Part No. 00403-90010
• New or Revised Item
ERRATA:
Page 1-1, Table 1-1. Add the following to Model 403B and 403B Option 01 accuracy specification tables: "On the 300 V range, accuracy is not specified above
100 kHz."
Page 5-3, Para.5-15d. Change frequency from 5 Hz to
10 Hz.
Page 5-3, Paragraph 5-16(b). Change the last sentence to agree with the following: "The 403B meter deflection on any range should be less than 3% of full scale with battery operation and less than 8% of full scale with an ac source
(see Paragraph 3~)."
Page 5-9/5-10, Figure 5~.
Add ferrite bead, A 1 L1, to emitter of A 1Q6.
Change adjustment designation under A3C3 to "CAL 30 V,
300 KC."
Section' V, Performance Check Test Card. Number 4,
Noise Check results should read
"<
3% of full scale deflection on any range" for Battery Operation and
"<
8% of full scale deflection on any range" for ac operation.
Page 6-2, Table 6-1.
Add Ref.
Desig. L1, Part No.
9170~016, ferrite bead. Add L 1 to 403B-65A assembly.
Page 6-3, Table 6-1. Change description of S1 to read
"Assembly RANGE switch: 3 sect 12 pos; includes Cl,
C6, A3 and R 17 thru R23."
Page 6-4, Table 6-1. Operating and Service Manual is Part
No. 00403-90010.
Page 6-5, Table 6-2. Item No.8 is Part No. 5000.a569
Cover Assy: Bottom.
I
Page 5-9/5-10, Figure 5-8. Add Component Locator for
Al board as in Figure 1.
CHANGE NO.1: Applies to all instruments.
Page 6-2, Table 6-1. Change Part No. of CR3, CR4 to
1901.0518. Previous Part is hard to obtain; this is the preferred replacement.
CHANGE NO.2: Applies to 4038 Serial No's.
0986A 15246 and greater and 4038 Option 01 Serial No's.
0986A 15346 and greater.
Page 6-3, Table 6-1. Change R13 to Part No. 0687-2231,
22 K ± 10% 1/2 W. This change desensitizes Ql from bias fluctuation.
Figure 1. Component Locator..
tip
AI
Port No. 403 B -65A
Rev.B
NOTE: See Figure 5-1 for additional location information.
17 September 1975
~
--1
Cll
M-
e
- R I 6 -
-RI~-
- R I 4 -
- R I I -
RI3 Rll
I
I I
(0
- C R -
- C R l -
- R 9 - -
- R 8 - -
- R 1 -
-RIO-
- ' 8 -
- R 6 -
- C 1 -
ClO
I
-CR4-
-»~,E~ ~
• • T .
- R 4 0 -
I
I
I
R3l
I
I
CI9
-CR3e e
I
-R31~
Rl8
~ CI~
I
~16) 'Ri
I
I
R46
I
CAL 1 mV
2 MHz
~----~- METER CAL
1 mV RANGE
9
I
R34
-R24-
~
~
"21
I
1
"'!.l
"----"
~.
LOOP GAIN
ADJ
Supplement A for 00403-900 10
00403-90010
HEWLETT
iii
PACKARD
PRINTED IN U.S.A.
Lataa