ENDRESS+HAUSER Proline Promag 50P Technical information

Add to My manuals
40 Pages

advertisement

ENDRESS+HAUSER Proline Promag 50P Technical information | Manualzz
Technical Information
Proline Promag 50P, 53P
Electromagnetic Flow Measuring System
Flow rate measurement in chemical or process applications
Application
Electromagnetic flowmeter for bidirectional
measurement of liquids with a minimum conductivity
of ≥ 5 μS/cm:
• Acids and caustic solutions
• Paints
• Pastes, mashes
• Water, wastewater etc.
• Flow measurement up to 44,000 GPM
(9600 m3/h)
• Fluid temperature up to 356°F (180°C)
• Process pressures up to 580 psi (40 bar)
• Fitting lengths to DVGW/ISO
Application-specific lining materials:
• PTFE und PFA
Approvals for hazardous area:
• ATEX, FM, CSA, TIIS
Connection to process control system:
• HART®, Profibus® DP/PA, FOUNDATION™
Fieldbus, MODBUS® RS485
TI047D/24/ae/07.08
Your benefits
Promag measuring devices offer you cost-effective flow
measurement with a high degree of accuracy for a wide
range of process conditions.
The Proline transmitter concept comprises:
• Modular device and operating concept resulting in a
higher degree of efficiency
• Software options for batching, electrode cleaning and
for measuring pulsating flow.
• Uniform operating concept
The tried-and-tested Promag sensors offer:
• No pressure loss
• Not sensitive to vibrations
• Simple installation and commissioning
Proline Promag 50P, 53P
Function and system design
Measuring principle
Faraday’s law of induction states that a voltage is induced in a conductor moving in a magnetic field.
In electromagnetic measuring, the flowing medium corresponds to the moving conductor. The induced voltage
is proportional to the flow velocity and is detected by two measuring electrodes and transmitted to the amplifier. Flow volume is computed on the basis of the pipe's diameter. The constant magnetic field is generated by
a switched direct current of alternating polarity.
Ue
I
V
B
I
L
A0003191
Ue = B · L · v
Q=A·v
Ue = induced voltage
B = magnetic induction (magnetic field)
L = electrode gap
v = flow velocity
Q = volume flow
A = pipe cross-section
I = current strength
Measuring system
The measuring system consists of a transmitter and a sensor.
Two versions are available:
• Compact version: transmitter and sensor form a single mechanical unit.
• Remote version: transmitter and sensor are installed separately.
Transmitter:
The Promag 50/53 transmitters incorporate a high impedance amplifier of 1 x 1012 ohms or greater.
• Promag 50 (user interface with push buttons for operation, two-line display)
• Promag 53 (“Touch Control” without opening the housing, four-line display)
Sensor:
• 1/2" to 24" (DN 15 to 600)
Input
Measured variable
Flow rate (proportional to induced voltage)
Measuring range
Typically v = 0.033 to 33 ft/s (0.01 to 10 m/s) with the specified measuring accuracy
Operable flow range
Over 1000 : 1
Input signal
Status input (auxiliary input):
U = 3 to 30 V DC, Ri = 5 kΩ, galvanically isolated.
Configurable for: totalizer(s) reset, measured value suppression, error-message reset.
Status input (auxiliary input) with PROFIBUS DP and MODBUS RS485:
U = 3 to 30 V DC, Ri = 3 kΩ, galvanically isolated
2
Endress+Hauser
Proline Promag 50P, 53P
Switching level: 3 to 30 V DC, independent of polarity
Configurable for: totalizer(s) reset, measured value suppression, error-message reset,
batching start/stop (optional), batch totalizer reset (optional)
Current input (for Promag 53 only):
Active/passive selectable, galvanically isolated, full scale value selectable, resolution: 3 μA,
temperature coefficient: typ. 0.003% o.r./°F (0.005% o.r./°C), o.r. = of reading
active: 4 to 20 mA, Ri ≤ 150 Ω, Uout = 24 V DC, short-circuit-proof
passive: 0/4 to 20 mA, Ri ≤ 150 Ω, Umax = 30 V DC
Output
Output signal
Promag 50
Current output:
active/passive selectable, galvanically isolated, time constant selectable (0.01 to 100 s),
full scale value selectable, temperature coefficient: typ. 0.003% o.r./°F (0.005% o.r./°C), o.r. = of reading,
resolution: 0.5 μA
• active: 0/4 to 20 mA, RL < 700 Ω (HART: RL ≥ 250 Ω)
• passive: 4 to 20 mA, operating voltage VS 18 to 30 V DC, Ri ≤ 150 Ω
Pulse/frequency output:
passive, open collector, 30 V DC, 250 mA, galvanically isolated.
• Frequency output: full scale frequency 2 to 1000 Hz (fmax = 1250 Hz), on/off ratio 1:1,
pulse width max. 10 s.
• Pulse output: pulse value and pulse polarity selectable, max. pulse width configurable
(0.5 to 2000 ms)
PROFIBUS DP interface:
• Transmission technology (Physical Layer): RS485 in accordance with ANSI/TIA/EIA-485-A: 1998,
galvanically isolated
• Profile version 3.0
• Data transmission rate: 9.6 kBaud to 12 MBaud
• Automatic data transmission rate recognition
• Function blocks: 1 x analog input, 3 x totalizer
• Output data: volume flow, totalizer
• Input data: positive zero return (ON/OFF), totalizer control, value for local display
• Cyclic data transmission compatible with previous model “Promag 33”
• Bus address adjustable via miniature switches or local display (optional) at the measuring device
Endress+Hauser
3
Proline Promag 50P, 53P
PROFIBUS PA interface:
• Transmission technology (Physical Layer): IEC 61158-2 (MBP), galvanically isolated
• Profile version 3.0
• Current consumption: 11 mA
• Permissible supply voltage: 9 to 32 V
• Bus connection with integrated reverse polarity protection
• Error current FDE (Fault Disconnection Electronic): 0 mA
• Function blocks: 1 x analog input, 1 x totalizer
• Output data: volume flow, totalizer
• Input data: positive zero return (ON/OFF), control totalizer, value for local display
• Cyclic data transmission compatible with previous model “Promag 33”
• Bus address adjustable via miniature switches or local display (optional) at the measuring device
Promag 53
Current output:
active/passive selectable, galvanically isolated, time constant selectable (0.01 to 100 s),
full scale value selectable, temperature coefficient: typically 0.005% o.r./°C (o.r. = of reading),
resolution: 0.5 μA
• active: 0/4 to 20 mA, RL < 700 Ω (HART: RL ≥ 250 Ω)
• passive: 4 to 20 mA, operating voltage VS 18 to 30 V DC, Ri ≤ 150 Ω
Pulse/frequency output:
active/passive selectable, galvanically isolated (Ex i version: only passive)
• active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100 Ω
• passive: open collector, 30 V DC, 250 mA
• Frequency output: full scale frequency 2 to 10000 Hz (fmax = 12500 Hz), EEx-ia: 2 to 5000 Hz;
on/off ratio 1:1; pulse width max. 10 s.
• Pulse output: pulse value and pulse polarity adjustable, pulse width configurable
(0.05 to 2000 ms)
PROFIBUS DP interface:
• Transmission technology (Physical Layer): RS485 in accordance with ANSI/TIA/EIA-485-A:
1998, galvanically isolated
• Profile version 3.0
• Data transmission rate: 9.6 kBaud to 12 MBaud
• Automatic data transmission rate recognition
• Function blocks: 2 x analog input, 3 x totalizer
• Output data: volume flow, calculated mass flow, totalizer 1 to 3
• Input data: positive zero return (ON/OFF), totalizer control, value for local display
• Cyclic data transmission compatible with previous model “Promag 33”
• Bus address adjustable via miniature switches or local display (optional) at the measuring device
• Available output combination → Page 7 ff.
PROFIBUS PA interface:
• Transmission technology (Physical Layer): IEC 61158-2 (MBP), galvanically isolated
• Profile version 3.0
• Current consumption: 11 mA
• Permissible supply voltage: 9 to 32 V
• Bus connection with integrated reverse polarity protection
• Error current FDE (Fault Disconnection Electronic): 0 mA
• Function blocks: 2 x analog input, 3 x totalizer
• Output data: volume flow, calculated mass flow, totalizer 1 to 3
• Input data: positive zero return (ON/OFF), totalizer control, value for local display
• Cyclic data transmission compatible with previous model “Promag 33”
• Bus address adjustable via miniature switches or local display (optional) at the measuring device
4
Endress+Hauser
Proline Promag 50P, 53P
MODBUS interface:
• Transmission technology (Physical Layer): RS485 in accordance with ANSI/TIA/EIA-485-A:
1998, galvanically isolated
• MODBUS device type: Slave
• Address range: 1 to 247
• Bus address adjustable via miniature switches or local display (optional) at the measuring device
• Supported MODBUS function codes: 03, 04, 06, 08, 16, 23
• Broadcast: supported with the function codes 06, 16, 23
• Transmission mode: RTU oder ASCII
• Supported baudrate: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 Baud
• Response time:
Direct data access = typically 25 to 50 ms
Auto-scan buffer (data range) = typically 3 to 5 ms
• Available output combination → Page 7 ff.
FOUNDATION Fieldbus interface:
• FOUNDATION Fieldbus H1
• Transmission technology (Physical Layer): IEC 61158-2 (MBP), galvanically isolated
• ITK version 4.01
• Current consumption: 12 mA
• Error current FDE (Fault Disconnection Electronic): 0 mA
• Bus connection with integrated reverse polarity protection
• Function blocks: 5 x analog input, 1 x discrete output, 1 x PID
• Output data: volume flow, calculated mass flow, temperature, totalizer 1 to 3
• Input data: positive zero return (ON/OFF), reset totalizer
• Link Master (LM) functionality is supported
Current output → failure response selectable (e.g. in accord. with NAMUR Recom. NE 43)
Pulse/frequency output → failure response selectable
Status output (Promag 50) → non-conductive by fault or power supply failure
Relay output (Promag 53) → de-energized by fault or power supply failure
Signal on alarm
•
•
•
•
Load
See “Output signal”
Switching output
Status output (Promag 50, Promag 53):
Open collector, max. 30 V DC / 250 mA, galvanically isolated.
Configurable for: error messages, Empty Pipe Detection (EPD), flow direction, limit values.
Relay outputs (Promag 53):
Normally closed (NC or break) or normally open (NO or make) contacts available
(default: relay 1 = NO, relay 2 = NC)
max. 30 V / 0.5 A AC; 60 V / 0.1 A DC, galvanically isolated.
Configurable for: error messages, Empty Pipe Detection (EPD), flow direction, limit values, batching contacts.
Low flow cutoff
Switch points for low flow cutoff are selectable
Galvanic isolation
All circuits for inputs, outputs, and power supply are galvanically isolated from each other.
Endress+Hauser
5
Proline Promag 50P, 53P
Power supply
Electrical connection
Measuring unit
A
B
d
g
b
a
a
b d/(g) (d)
PROFIBUS PA*
FOUNDATION Fieldbus*
HART*
f
– 27
+ 26
– 25
+ 24
– 23
+ 22
– 21
+ 20
N (L-) 2
L1 (L+)1
d
e
c
PA(–)/FF(–)
PA(+)/FF(+)
–
+
f
–
+
–
+
27
26
25
24
23
22
21
20
N (L-) 2
L1 (L+) 1
b
N (L-) 2
L1 (L+) 1
d
e
g
c
b
Cable cross section max. 16 AWG (2.5 mm2)
e
c
b
PROFIBUS DP**
MODBUS RS485**
PROFIBUS DP*
A (RxD/TxD-N) 27
B (RxD/TxD-P) 26
– 25
+ 24
f
– 23
+ 22
– 21
+ 20
d
A (RxD/TxD-N)
B (RxD/TxD-P)
–
+
f
–
+
–
+
27
26
25
24
23
22
21
20
N (L-) 2
L1 (L+) 1
d
e
g
c
b
a0002441
A View A (field housing)
B View B (wall-mount housing)
*)
**)
a
b
c
d
e
f
g
6
not changeable communication board
changeable communication board
Cover of the connection compartment
Cable for power supply: 85 to 260 V AC, 20 to 55 V AC,16 to 62 V DC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 2: N for AC, L- for DC
Ground terminal for protective conductor
Signal cable: see Terminal assignment → Page 7 ff.
Fieldbus cable:
Terminal No. 26: DP (A) / PA (+) / FF (+) / MODBUS RS485 (A) / (PA, FF: with reverse polarity protection)
Terminal No. 27: DP (B) / PA (-) / FF (-) / MODBUS RS485 (B) / (PA, FF: with reverse polarity protection)
Ground terminal for signal-cable shield / Fieldbus cable / RS485 line
Service connector for connecting service interface FXA 193 (Fieldcheck, Tof Tool - Fieldtool Package)
Signal cable: see Terminal assignment → Page 7 ff.
Cable for external termination (only for PROFIBUS DP with fixed assignment communication board):
Terminal No. 24: +5 V
Terminal No. 25: DGND
Endress+Hauser
Proline Promag 50P, 53P
Terminal assignment, Promag 50
Terminal No. (inputs / outputs)
Order variant
20 (+) / 21 (–)
22 (+) / 23 (–)
24 (+) / 25 (–)
26 (+) / 27 (–)
50***-***********W
−
−
−
Current output
HART
50***-***********A
−
−
Frequency output
Current output
HART
50***-***********D
Status input
Status output
Frequency output
Current output
HART
50***-***********H
–
–
–
50***-***********J
–
–
+5 V (external
termination)
PROFIBUS DP
50***-***********S
–
–
Frequency output
Ex i, passive
Current output Ex i
active, HART
50***-***********T
–
–
Frequency output
Ex i, passive
Current output Ex i
passive, HART
PROFIBUS PA
Ground connection, power supply → Page 6
Terminal assignment, Promag 53
The inputs and outputs on the communication board can be either permanently assigned or variable, depending
on the version ordered (see table). Replacements for modules which are defective or which have to be replaced
can be ordered as accessories.
Terminal No. (inputs / outputs)
Order variant
20 (+) / 21 (–)
22 (+) / 23 (–)
24 (+) / 25 (–)
26 (+) / 27 (–)
Fixed communication boards (fixed assignment)
53***-***********A
−
−
Frequency output
Current output
HART
53***-***********B
Relay output 2
Relay output 1
Frequency output
Current output
HART
53***-***********F
–
–
–
PROFIBUS PA
IS
53***-***********G
–
–
–
FOUNDATION
Fieldbus, IS
53***-***********H
–
–
–
53***-***********J
–
–
–
53***-***********K
−
−
−
53***-***********Q
−
−
Status input
53***-***********S
−
−
Frequency output
IS
Current output IS
active, HART
53***-***********T
−
−
Frequency output
IS
Current output IS
passive, HART
53***-***********C
Relay output 2
Relay output 1
Frequency output
Current output
HART
53***-***********D
Status input
Relay output
Frequency output
Current output
HART
53***-***********L
Status input
Relay output 2
Relay output 1
Current output
HART
PROFIBUS PA
PROFIBUS DP
FOUNDATION
Fieldbus
MODBUS RS485
Flexible communication boards
Endress+Hauser
7
Proline Promag 50P, 53P
Terminal No. (inputs / outputs)
Order variant
20 (+) / 21 (–)
22 (+) / 23 (–)
24 (+) / 25 (–)
26 (+) / 27 (–)
53***-***********M
Status input
Frequency output
Frequency output
Current output
HART
53***-***********N
Current output
Frequency output
Status input
MODBUS RS485
53***-***********P
Current output
Frequency output
Status input
PROFIBUS DP
53***-***********V
Relay output 2
Relay output 1
Status input
PROFIBUS DP
53***-***********2
Relay output
Current output
Frequency output
Current output
HART
53***-***********4
Current input
Relay output
Frequency output
Current output
HART
53***-***********5
Status input
Current input
Frequency output
Current output
HART
53***-***********7
Relay output 2
Relay output 1
Status input
MODBUS RS485
Ground connection, power supply → Page 6
Electrical connection remote
version
Coil circuit
Electrode circuit
Meas. signal Pipe EPD
S1 E1 E2 S2 GND E S
n.c.
n.c.
5
7
2
42 41
1
4 37 36
yellow
8
green
brown
7
white
5
6
n.c.
4 37
42 41
F06-5xFxxxxx-04-xx-xx-de-001
n.c. = isolated cable shields, not connected
Cable entry
Power-supply and signal cables (inputs/outputs):
• Cable entry M20 x 1.5 (8 to 12 mm)
• Sensor cable entry for armoured cables M20 x 1.5 (9.5 to 16 mm)
• Threads for cable entries 1/2" NPT, G 1/2"
Connecting cable for remote version:
• Cable entry M20 x 1.5 (8 to 12 mm)
• Sensor cable entry for armoured cables M20 x 1.5 (9.5 to 16 mm)
• Threads for cable entries 1/2" NPT, G 1/2"
8
Endress+Hauser
Proline Promag 50P, 53P
Cable specifications
remote version
Coil cable:
• 2 x 18 AWG (0.75 mm2) PVC cable with common, braided copper shield (Ø approx. 7 mm)
• Conductor resistance: ≤ 37 Ω/km
• Capacitance: core/core, shield grounded: ≤ 120 pF/m
• Permanent operating temperature: -4 to +180°F (–20 to +80°C)
• Cable cross-section: max. 16 AWG (2.5 mm2)
Signal cable:
• 3 x 20 AWG (0.38 mm2) PVC cable with common, braided copper shield (Ø approx. 0.28" / 7 mm) and
individually shielded cores
• With Empty Pipe Detection (EPD): 4 x 20 AWG (0.38 mm2) PVC cable with common, braided copper shield
(Ø approx. 0.28" / 7 mm) and individually shielded cores.
• Conductor resistance: ≤ 50 Ω/km
• Capacitance: core/shield: ≤ 420 pF/m
• Permanent operating temperature: -4 to +180°F (–20 to +80°C)
• Cable cross-section: max. 16 AWG (2.5 mm2)
1
2
3
4
5
6
7
a
b
a = signal cable, b = coil current cable (cross-section: max. 16 AWG / 2.5 mm2)
1 = core, 2 = core insulation, 3 = core shield, 4 = core jacket, 5 = core strengthening,
6 = cable shield, 7 = outer jacket
A0003194
Optionally, Endress+Hauser also supplies reinforced connecting cables with an additional, metal strenghtening
braid. We recommend such cables for the following cases:
• Cables laid underground
• Danger of rodent attack
• Device used with ingress protection NEMA 6P (IP 68)
Operation in zones of severe electrical interference:
The measuring device complies with the general safety requirements in accordance with EN 61010,
the EMC requirements of EN 61326/A1, and NAMUR recommendation NE 21.
Caution!
Grounding is by means of the ground terminals provided for that purpose inside the connection housing.
Keep the stripped and twisted lengths of cable shield to the terminals as short as possible.
Endress+Hauser
9
Proline Promag 50P, 53P
Supply voltage
85 to 260 V AC, 45 to 65 Hz
20 to 55 V AC, 45 to 65 Hz
16 to 62 V DC
PROFIBUS PA and FOUNDATION Fieldbus
Non-hazardous: 9 to 32 V DC
Intrinsically safe: 9 to 24 V DC
Explosion proof: 9 to 32 V DC
Power consumption
AC: <15 VA (including sensor)
DC: <15 W (including sensor)
Switch-on current:
• max. 13.5 A (< 50 ms) at 24 V DC
• max. 3 A (< 5 ms) at 260 V AC
Power supply failure
Lasting min. 1 power cycle:
• EEPROM or T-DAT (Promag 53 only) retain the measuring system data in the event of a power supply failure
• S-DAT: exchangeable data storage chip which stores the data of the sensor (nominal diameter, serial number,
calibration factor, zero point, etc.)
Potential equalisation
Standard case
Perfect measurement is only ensured when the medium and the sensor have the same electrical potential. Most
Promag sensors have a standard installed reference electrode which guarantees the required connection. This
usually means that additional potential matching measures are unnecessary.
Note!
For installation in metal pipes, it is advisable to connect the ground terminal of the transmitter housing to the
piping. Also, observe company-internal grounding guidelines.
A0004375
Caution!
For sensors without reference electrodes or without metal process terminals, carry out potential matching as
per the instructions for special cases described below. These special measures are particularly important when
standard grounding practice cannot be ensured or extremely strong matching currents are expected.
10
Endress+Hauser
Proline Promag 50P, 53P
Metal, ungrounded piping
In order to prevent outside influences on measurement, it is advisable to use ground cables to connect each
sensor flange to its corresponding pipe flange and ground the flanges. Connect the transmitter or sensor connection housing, as applicable, to ground potential by means of the ground terminal provided for the purpose.
Caution!
Also, observe company-internal grounding guidelines.
Note!
The ground cable for flange-to-flange connections can be ordered separately as an accessory from
Endress+Hauser.
• Sensor sizes ≤ 12" (DN ≤ 300): The ground cable is in direct connection with the conductive flange coating
and is secured by the flange screws.
• Sensor sizes ≥ 14" (DN ≥ 350): The ground cable connects directly to the metal transport bracket.
9 AWG (6 mm² Cu)
Pipe sizes 12”
Pipe sizes 14”
A0004376
Endress+Hauser
11
Proline Promag 50P, 53P
Plastic pipes and isolating lined pipes
Normally, potential is matched using the reference electrodes in the measuring tube. However, in exceptional
cases it is possible that, due to the grounding plan of a system, large matching currents flow over the reference
electrodes. This can lead to destruction of the sensor, e.g. through electrochemical decomposition of the electrodes. In such cases, e.g. for fibre-glass or PVC piping, it is recommended that you use additional ground disks
for potential matching.
When using ground disks, note the following points:
• Ground disks (1/2" to 12" / DN 15 to 300) can be ordered separately from Endress+Hauser as an accessory.
• Ground disks (incl. seals) increase the installation length. You can find the dimensions of ground disks on
Page 31.
Caution!
• Risk of damage from electrochemical corrosion. Note the electrochemical insulation rating, if the ground
disks and measuring electrodes are made of different materials.
• Also, observe company-internal grounding guidelines.
9 AWG (6 mm² Cu)
A0004377
Pipes with cathodic protection
In such cases, install the measuring instrument without potential in the piping:
• When installing the measuring device, make sure that there is an electrical connection between the two
piping runs (copper wire, 9 AWG / 6 mm2).
• Make sure that the installation materials do not establish a conductive connection to the measuring device
and that the installation materials withstand the tightening torques applied when the threaded fasteners are
tightened.
• Also comply with the regulations applicable to potential-free installation.
2
2
1
9 AWG Cu
A0004378
1 = isolation transformer, 2 = electrically isolated
12
Endress+Hauser
Proline Promag 50P, 53P
Performance characteristics
Reference operating
conditions
To DIN EN 29104 and VDI/VDE 2641:
• Medium temperature: +28°C ± 2 K
• Ambient temperature: +22°C ± 2 K
• Warm-up period: 30 minutes
Installation:
• Inlet run >10 x Dia. (DN)
• Outlet run > 5 x Dia. (DN)
• Sensor and transmitter grounded.
• Sensor centered relative to the pipe.
Maximum measured error
Promag 50:
Pulse output: ± 0.5% o.r. ± 1 mm/s (o.r. = of reading)
Current output: plus typically ± 5 μA
Promag 53:
Pulse output: ± 0.2% o.r. ± 2 mm/s (o.r. = of reading)
Current output: plus typically ± 5 μA
Supply voltage fluctuations have no effect within the specified range.
[%]
2.5
2.0
0.5 %
1.5
0.2 %
1.0
0.5
0
0
3
(1)
6
(2)
13
(4)
20
(6)
26
(8)
33
(10)
v in ft/s [m/s]
F06-5xxxxxxx-05-xx-xx-xx-000
Max. measured error in % of reading
Repeatability
max. ± 0.1% o.r. ± 0.5 mm/s (o.r. = of reading)
Operating conditions
Installation conditions
Installation instructions
Endress+Hauser
Mounting location
Correct measuring is possible only if the pipe is full. Avoid the following locations:
• Highest point of a pipeline. Risk of air accumulating.
• Directly upstream of a free pipe outlet in a vertical pipe.
13
Proline Promag 50P, 53P
h 2 x Dia
F06-5xxxxxxx-11-00-00-xx-000
Installation of pumps
Do not install the sensor on the intake side of a pump. This precaution is to avoid low pressure and the consequent risk of damage to the lining of the measuring tube. Information on the lining's resistance to partial vacuum can be found on Page 21.
It might be necessary to install pulse dampers in systems incorporating reciprocating, diaphragm or peristaltic
pumps. Information on the measuring system's resistance to vibration and shock can be found on Page 20.
F06-5xxxxxxx-11-00-00-xx-001
14
Endress+Hauser
Proline Promag 50P, 53P
Partially filled pipes
Partially filled pipes with gradients necessitate a drain-type configuration. The Empty Pipe Detection (EPD)
function offers additional protection by detecting empty or partially filled pipes.
Caution!
Risk of solids accumulating. Do not install the sensor at the lowest point in the drain. It is advisable to install a
cleaning valve.
2 x Dia.
5 x Dia.
F06-5xxxxxxx-11-00-00-xx-002
Vertical pipes
Install a siphon (b) or a vent valve (a) downstream of the sensor in vertical pipes longer than 5 meters. This
precaution is to avoid low pressure and the consequent risk of damage to the lining of the measuring tube.
These measures also prevent the system losing prime, which could cause air inclusions. Information on the lining’s resistance to partial vacuum can be found on Page 21.
a
b
> 15 ft (5 m)
F06-5xxxxxxx-11-00-00-xx-003
a = vent valve, b = siphon
Endress+Hauser
15
Proline Promag 50P, 53P
Orientation
An optimum orientation helps avoid gas and air accumulations and deposits in the measuring tube. Promag,
nevertheless, supplies a range of options and accessories for correct measuring of problematic mediums:
• Electrode Cleaning Circuitry (ECC) to remove electrically conductive deposits in the measuring tube, e.g. in
accretive mediums.
• Empty Pipe Detection (EPD) for recognition of partially filled measuring tubes, or for degassing mediums or
for applications with fluctuating process pressure.
Vertical orientation:
This orientation is ideal for self-emptying piping systems and for use in conjunction with Empty Pipe Detection.
F06-5xxxxxxx-11-00-00-xx-004
Horizontal orientation:
The measuring electrode-plane should be horizontal. This prevents brief insulation of the two electrodes by
entrained air bubbles.
Caution!
Empty Pipe Detection functions correctly only when the measuring device is installed horizontally and the
transmitter housing is facing upward. Otherwise there is no guarantee that Empty Pipe Detection will respond
if the measuring tube is only partially filled or empty.
F06-5xxxxxxx-11-00-xx-xx-000
1 = EPD electrode (Empty Pipe Detection)
2 = Measuring electrodes (signal detection)
3 = Reference electrode (potential equalisation)
16
Endress+Hauser
Proline Promag 50P, 53P
Vibrations
Secure the piping and the sensor if vibration is severe.
Caution!
It is advisable to install sensor and transmitter separately if vibration is excessively severe.
Information on resistance to vibration and shock can be found on Page 20.
> 30 ft ( 10 m)
F06-5xxxxxxx-11-00-00-xx-006
Foundations, supports
If the nominal diameter is ≥ 14" (DN ≥ 350), mount the transmitter on a foundation of adequate loadbearing strength.
Caution!
Do not allow the casing to take the weight of the sensor. This would buckle the casing and
damage the internal magnetic coils.
F06-5xFxxxxx-11-05-xx-xx-000
Endress+Hauser
17
Proline Promag 50P, 53P
Inlet and outlet runs
If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc. Compliance with the
following requirements for the inlet and outlet runs is necessary in order to ensure measuring accuracy:
• Inlet run ≥ 5 x Dia
• Outlet run ≥ 2 x Dia
5 x Dia.
2 x Dia.
F06-5xxxxxxx-11-00-00-xx-005
Adapters
Suitable adapters to DIN EN 545 (double-flange junction sections) can be used to install the sensor in largerdiameter pipes. The resultant increase in the rate of flow improves measuring accuracy with very slow-moving
fluids.
The nomogram shown here can be used to calculate the pressure loss caused by reducers and expanders.
The nomogram applies only to fluids of viscosity similar to water:
1. Calculate the ratio of the diameters d/D.
2. From the nomogram read off the pressure loss as a function of flow velocity (downstream from the reduction) and the d/D ratio.
psi [mbar]
1.45 (100)
26 ft/s
8 m/s
23 ft/s
7 m/s
20 ft/s
6 m/s
0.14 (10)
16 ft/s
5 m/s
13 ft/s
4 m/s
max. 8 °
d
10 ft/s
3 m/s
D
6 ft/s
2 m/s
0.014 (1)
3 ft/s
1 m/s
0.5
0.6
0.7
0.8
0.9
d/D
F06-5xFxxxxx-05-xx-xx-xx-000
18
Endress+Hauser
Proline Promag 50P, 53P
Length of connecting cable
Permissible cable length Lmax depends on the conductivity of the medium. A minimum conductivity of
20 μS/cm is required for measuring demineralized water.
(S/cm]
200
L
max
100
5
33
(10)
330
(100)
L
650
(200)
ft
[m]
max
F06-5xxxxxxx-05-xx-xx-xx-006
Gray shaded area = permissible range for medium conductivity
Lmax = length of connecting cable in [m]
Medium conductivity in [μS/cm]
In order to ensure measuring accuracy, moreover, comply with the following instructions when installing the
remote version:
• Secure the cable run or route the cable in a conduit. Movement of the cable can falsify the measuring signal,
particularly if the conductivity of the medium is low.
• Route the cable well clear of electrical machines and switching elements.
• Ensure potential equalisation between sensor and transmitter, if necessary.
Environment
Ambient temperature
Transmitter:
• Standard: -5 to +140°F (–20 to +60°C)
• Optional: -40 to +140°F (–40 to +60°C)
Note!
At ambient temperatures below -5°F (–20°C) the readability of the display may be impaired.
Sensor:
• Flange material carbon steel: 14 to +140°F (–10 to +60°C)
• Flange material stainless steel: -40 to +140°F (–40 to +60°C)
Caution!
It is not allowed to use the device beyond the min. and max. lining specified temperature values
(→ “Medium temperature range”).
Note the following points:
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• If both fluid and ambient temperatures are high, install the transmitter at a remote location from the sensor
(→ “Medium temperature range”).
Note!
HE or Harsh Environment Protection for sensor:
Provides enchanced moisture ingress protection for humid ambient environments. Apply in applications with
severe temperature cycling and humid environments. Refer to ordering information, page 38.
Storage temperature
Endress+Hauser
• The storage temperature corresponds to the ambient temperature range of the transmitter and sensor
(see “Ambient temperature”).
• The measuring device must be protected against direct sunlight during storage in order to avoid unacceptably
high surface temperatures.
• Choose a storage location where moisture does not collect in the measuring device. This will help prevent
fungus and bacteria infestation which can damage the liner.
• Do not remove the protective plates or caps on the process connections until the device is ready to install.
This is particularly important in the case of sensors with PTFE linings.
19
Proline Promag 50P, 53P
Degree of protection
• Standard: NEMA 4X (IP 67) for transmitter and sensor
• Optional: NEMA 6P (IP 68) for sensor in remote version
Shock and vibration resistance
Acceleration up to 2 g by analogy with IEC 60068-2-6
(high-temperature version: no data available)
Electromagnetic compatibility
(EMC)
To EN 61326/A1 and NAMUR recommendation NE 21
Process conditions
Medium temperature range
The permissible medium temperature depends on the measuring-tube lining:
• -40 to +265°F (−40 to +130°C) for PTFE (1/2" to 24" / DN 15 to 600), for restrictions → refer to
diagrams
• -4 to + 355°F (−20 to +180°C for PFA (1" to 8" / DN 25 to 200), for restrictions → refer to diagrams
Compact version (PFA and PTFE lining)
TA = ambient temperature, TF = fluid temperature, HT = high-temperature version, with insulation
m = temperature range from 14 to -40°F (-10°C to -40°C) only applies for flanges made of stainless steel
HT
TA °F [°C]
140 (60)
104 (40)
68 (20)
32 (0)
PF A
-4 (-20)
m
PTFE
-40 (-40)
-40 -4 32 68 104 140 176 212 248 284 320 356
(-40) (-20) (0) (20) (40) (60) (80) (100) (120) (140) (160) (180)
TF °F [°C]
A0002660
Remote version (PFA and PTFE lining)
TA = ambient temperature, TF = fluid temperature, HT = high-temperature version, with insulation
m = temperature range from 14 to -40°F (-10°C to -40°C) only applies for flanges made of stainless steel
HT
TA °F [°C]
140 (60)
104 (40)
68 (20)
32 (0)
-4 (-20)
-40 (-40)
PF A
PTFE
m
-40 -4 32 68 104 140 176 212 248 284 320 356 TF
(-40) (-20) (0) (20) (40) (60) (80) (100) (120) (140) (160) (180)
°F [°C]
A0002671
20
Endress+Hauser
Proline Promag 50P, 53P
Conductivity
Minimum conductivity:
≥ 5 μS/cm for fluids generally
≥ 20 μS/cm for demineralised water
Note that in the case of the remote version, the minimum conductivity is also influenced by the length of the
connecting cable → see “Length of connecting cable”
Medium pressure range
(nominal pressure)
EN 1092-1 (DIN 2501):
PN 10 (DN 200 to 600)
PN 16 (DN 65 to 600)
PN 25 (DN 200 to 600)
PN 40 (DN 15 to 150)
ANSI B16.5:
Class 150 (1/2" to 24")
Class 300 (1/2" to 6")
JIS B2238:
10K (DN 50 to 300)
20K (DN 15 to 300)
AS2129:
Table E (DN 25, 50)
AS4087:
Cl. 14 (DN 50)
Pressure tightness
(liner)
Nominal
diameter
[mm]
[inch]
15
1/2"
25
1"
Measuring
tube
lining
Resistance to partial vacuum of measuring tube lining
Limit values for abs. pressure psi [mbar] at various fluid temperatures
75°F
(25°C)
175°F
(80°C)
212°F
(100°C)
265°F
(130°C)
300°F
(150°C)
355°F
(180°C)
PTFE
0
0
0
1.45 (100)
−
−
PTFE / PFA
0/0
0/0
0/0
1.45 (100) / 0
−/0
−/0
32
−
PTFE / PFA
0/0
0/0
0/0
1.45 (100) / 0
−/0
−/0
40
1 1/2"
PTFE / PFA
0/0
0/0
0/0
1.45 (100) / 0
−/0
−/0
50
2"
PTFE / PFA
0/0
0/0
0/0
1.45 (100) / 0
−/0
−/0
65
−
PTFE / PFA
0/0
*
0.6 (40) / 0
1.9 (130) / 0
−/0
−/0
80
3"
PTFE / PFA
0/0
*
0.6 (40) / 0
1.9 (130) / 0
−/0
−/0
100
4"
PTFE / PFA
0/0
*
1.9 (135) / 0
2.5 (170) / 0
−/0
−/0
125
−
PTFE / PFA
1.9 (135) / 0
*
3.5 (240) / 0
5.6 (385) / 0
−/0
−/0
150
6"
PTFE / PFA
1.9 (135) / 0
*
3.5 (240) / 0
5.6 (385) / 0
−/0
−/0
200
8"
PTFE / PFA
2.9 (200) / 0
*
4.2 (290) / 0
5.9 (410) / 0
−/0
−/0
250
10"
PTFE
4.8 (330)
*
5.8 (400)
7.7 (530)
−
−
300
12"
PTFE
5.8 (400)
*
7.3 (500)
9.1 (630)
−
−
350
14"
PTFE
6.8 (470)
*
8.7 (600)
10.6 (730)
−
−
7.8 (540)
*
9.7 (670)
11.6 (800)
−
−
400
16"
PTFE
450
18"
PTFE
500
20"
PTFE
600
24"
PTFE
No vacuum is permissible!
* No value can be specified.
Endress+Hauser
21
Proline Promag 50P, 53P
Limiting flow
The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The optimum velocity
of flow is 6 to 10 ft/s (2 to 3 m/s). The velocity of flow (v), moreover, has to be matched to the physical properties of the medium:
• v < 6 ft/s (2 m/s): for abrasive mediums such as potter's clay, lime milk, ore slurry etc.
• v > 6 ft/s (2 m/s): for accretive mediums such as wastewater sludge, etc.
Flow characteristics (SI units)
Nominal
diameter
22
Recommended
flow rate
Min./max. full scale value
(v ~ 0.3 or 10 m/s)
Factory settings
Full scale value
(v ~ 2.5 m/s)
Pulse weighting
(~ 2 pulse/s)
Creepage
(v ~ 0.04 m/s)
[mm]
[inch]
15
1/2"
4…100
dm3/min
25
dm3/min
0.20
dm3
0.5
dm3/min
25
1"
9…300
dm3/min
75
dm3/min
0.50
dm3
1
dm3/min
32
1 1/4"
15…500
dm3/min
125
dm3/min
1.00
dm3
2
dm3/min
3
3
3
40
1 1/2"
25…700
dm /min
200
dm /min
1.50
dm
3
dm3/min
50
2"
35…1100
dm3/min
300
dm3/min
2.50
dm3
5
dm3/min
65
2 1/2"
60…2000
dm3/min
500
dm3/min
5.00
dm3
8
dm3/min
80
3"
90…3000
dm3/min
750
dm3/min
5.00
dm3
12
dm3/min
1200
dm3/min
10.00
dm3
20
dm3/min
1850
dm3/min
15.00
dm3
30
dm3/min
100
4"
145…4700
dm3/min
125
5"
220…7500
dm3/min
150
6"
20…600
m3/h
150
m3/h
0.025
m3
2.5
m3/h
200
8"
35…1100
m3/h
300
m3/h
0.05
m3
5.0
m3/h
500
m3/h
0.05
m3
7.5
m3/h
250
10"
55…1700
m3/h
300
12"
80…2400
m3/h
750
m3/h
0.10
m3
10
m3/h
350
14"
110…3300
m3/h
1000
m3/h
0.10
m3
15
m3/h
400
16"
140…4200
m3/h
1200
m3/h
0.15
m3
20
m3/h
3
3
3
25
m3/h
450
18"
180…5400
m /h
1500
m /h
0.25
m
500
20"
220…6600
m3/h
2000
m3/h
0.25
m3
30
m3/h
600
24"
310…9600
m3/h
2500
m3/h
0.30
m3
40
m3/h
Endress+Hauser
Proline Promag 50P, 53P
Flow characteristics (US units)
Nominal
diameter
Pressure loss
Endress+Hauser
Recommended
flow rate
Min./max. full scale value
(v ~ 1 or ~ 33 ft/s)
Factory settings
Full scale value
(v ~ 8 ft/s)
Pulse weighting
(~ 2 pulse/s)
Creepage
(v ~ 0.1 ft/s)
[inch]
[mm]
1/2"
15
1.0 to 27
gal/min
6
gal/min
0.05
gal
0.10
gal/min
1"
25
2.5 to 80
gal/min
18
gal/min
0.20
gal
0.25
gal/min
1 1/4"
32
4 to 130
gal/min
30
gal/min
0.20
gal
0.50
gal/min
1 1/2"
40
7 to 190
gal/min
50
gal/min
0.50
gal
0.75
gal/min
2"
50
10 to 300
gal/min
75
gal/min
0.50
gal
1.25
gal/min
2 1/2"
65
16 to 500
gal/min
130
gal/min
1
gal
2.0
gal/min
3"
80
24 to 800
gal/min
200
gal/min
2
gal
2.5
gal/min
4"
100
40 to 1250
gal/min
300
gal/min
2
gal
4.0
gal/min
5"
125
60 to 1950
gal/min
450
gal/min
5
gal
7.0
gal/min
6"
150
90 to 2650
gal/min
600
gal/min
5
gal
12
gal/min
8"
200
155 to 4850 gal/min
1200
gal/min
10
gal
15
gal/min
10"
250
250 to 7500 gal/min
1500
gal/min
15
gal
30
gal/min
12"
300
350 to 10,600
gal/min
2400
gal/min
25
gal
45
gal/min
14"
350
500 to 15,000
gal/min
3600
gal/min
30
gal
60
gal/min
16"
400
600 to 19,000
gal/min
4800
gal/min
50
gal
60
gal/min
18"
450
800 to 24,000
gal/min
6000
gal/min
50
gal
90
gal/min
20"
500
1000 to 30,000
gal/min
7500
gal/min
75
gal
120
gal/min
24"
600
1400 to 44,000
gal/min
10,500
gal/min
100
gal
180
gal/min
• No pressure loss if the sensor is installed in a pipe of the same nominal diameter.
• Pressure losses for configurations incorporating adapters to DIN EN 545 → Page 18
23
Proline Promag 50P, 53P
Measuring tube specifications
Nominal diameter
Pressure rating
[mm]
[inch]
EN (DIN)
[bar]
15
1/2"
25
1"
32
Inside diameter of
measuring tube
AS 2129
AS 4087
ANSI
[lbs]
JIS
with PFA
inches (mm)
with PTFE
inches (mm)
PN 40
−
−
Cl 150
20K
−
0.59 (15)
PN 40
Table E
−
Cl 150
20K
0.91 (23)
1.02 (26)
−
PN 40
−
−
−
20K
1.26 (32)
1.38 (35)
40
1 1/2"
PN 40
−
−
Cl 150
20K
1.42 (36)
1.61 (41)
50
2"
PN 40
Table E
Cl.14
Cl 150
10K
1.89 (48)
2.05 (52)
65
−
PN 16
−
−
−
10K
2.48 (63)
2.64 (67)
80
3"
PN 16
−
−
Cl 150
10K
2.95 (75)
3.15 (80)
100
4"
PN 16
−
−
Cl 150
10K
3.98 (101)
4.09 (104)
125
−
PN 16
−
−
−
10K
4.96 (126)
5.08 (129)
150
6"
PN 16
−
−
Cl 150
10K
6.06 (154)
6.14 (156)
200
8"
PN 10
−
−
Cl 150
10K
7.91 (201)
7.95 (202)
250
10"
PN 10
−
−
Cl 150
10K
−
10.1 (256)
300
12"
PN 10
−
−
Cl 150
10K
−
12.0 (306)
350
14"
PN 10
−
−
Cl 150
−
−
13.3 (337)
400
16"
PN 10
−
−
Cl 150
−
−
15.2 (387)
450
18"
PN 10
−
−
Cl 150
−
−
17.0 (432)
500
20"
PN 10
−
−
Cl 150
−
−
19.2 (487)
600
24"
PN 10
−
−
Cl 150
−
−
23.3 (593)
Mechanical construction
Design / dimensions
24
Dimensions: Wall-mount housing (non hazardous area and II3G / zone 2)
Endress+Hauser
Proline Promag 50P, 53P
-
Esc
6.27"
(159.5)
+
E
9.84"
(250)
3.56"
(90.5)
3.54"
1.77"
(90)
(45)
5.31"
(135)
8.46"
(215)
3.18"
(81)
> 1.97"
(50)
2.08"
(53)
3.18"
(81)
4.01"
(102)
2.08"
(53)
3.74"
3.20" (95)
(81.5)
7.55"
(192)
0.45"
(11.5)
0.45"
(11.5)
A0001150
Dimensions: Remote field housing (II2GD / zone 1)
10.4"
(265)
9.53" *
(242)
9.45”
(240)
8.54” *
(217)
~8.11”
(206)
7.32”
(186)
6.57"
(167)
13.9"
(355)
3.94"
(100)
7.40"
5.24" (188)
(133)
* Blind version
Æ 0.34"
(8.6 M8)
3.94"
(100)
4.84"
(123)
A0002128
Endress+Hauser
25
Proline Promag 50P, 53P
There is a separate mounting kit for the wall-mounted housing. It can be ordered from Endress+Hauser as an
accessory. The following installation variants are possible:
• Panel-mounted installation
• Pipe mounting
Panel-mounted installation
+0.02
8.27” -0.02
+0.5
(210 -0.5 )
+0.02
9.65” -0.02
(245
+0.5
-0.5
)
4.33”
(110)
A0001131
Pipe mounting
Ø 0.79” to 2.75”
(20 to 70)
~6.10”
(155)
A0001132
26
Endress+Hauser
Proline Promag 50P, 53P
Compact version ≤ 12" (DN ≤ 300)
7.68"
(195)
6.61"
(168)
8.94"
(227)
8.15" *
(207)
6.30"
(160)
Esc
-
+
E
B
A
C
E
L (DVGW)
K
F06-53Fxxxxx-06-00-xx-xx-000
DN
L
A
B
C
K
E
EN (DIN) / JIS /
AS*
[mm]
ANSI
[inch]
inches
(mm)
inches
(mm)
inches
(mm)
inches
(mm)
inches
(mm)
inches
(mm)
15
1/2"
7.87 (200)
13.3 (341)
10.1 (257)
3.31 (84)
4.72 (120)
3.7 (94)
25
1"
7.87 (200)
13.3 (341)
10.1 (257)
3.31 (84)
4.72 (120)
3.7 (94)
32
−
7.87 (200)
13.3 (341)
10.1 (257)
3.31 (84)
4.72 (120)
3.7 (94)
40
1 1/2"
7.87 (200)
13.3 (341)
10.1 (257)
3.31 (84)
4.72 (120)
3.7 (94)
50
2"
7.87 (200)
13.3 (341)
10.1 (257)
3.31 (84)
4.72 (120)
3.7 (94)
65
−
7.87 (200)
15.4 (391)
11.1 (282)
4.29 (109)
7.08 (180)
3.7 (94)
80
3"
7.87 (200)
15.4 (391)
11.1 (282)
4.29 (109)
7.08 (180)
3.7 (94)
100
4"
9.84 (250)
15.4 (391)
11.1 (282)
4.29 (109)
7.08 (180)
3.7 (94)
125
−
9.84 (250)
18.6 (472)
12.7 (322)
5.90 (150)
10.2 (260)
5.51 (140)
150
6"
11.8 (300)
18.6 (472)
12.7 (322)
5.90 (150)
10.2 (260)
5.51 (140)
200
8"
13.8 (350)
20.7 (527)
13.6 (347)
7.08 (180)
12.7 (324)
6.14 (156)
250
10"
17.7 (450)
22.7 (577)
14.6 (372)
8.07 (205)
15.7 (400)
6.14 (156)
300
12"
19.7 (500)
24.7 (627)
15.6 (397)
9.05 (230)
18.1 (460)
6.53 (166)
The fitting length (L) is always the same, regardless of the pressure rating.
* Only DN 25 and 50 are available if flanges according to AS are used.
High temperature version ≤ 12" (DN ≤ 300)
Esc
-
+
E
B1
4.33"
(110)
A1
F06-5xPxxxxx-06-00-00-xx-000
Measurement A1, B1 = Measurement A, B of the standard compact version plus 4.33" (110 mm)
Endress+Hauser
27
Proline Promag 50P, 53P
Remote version ≤ 12" (DN ≤ 300)
5.08"
(129)
6.73"
(171)
5.63"
(143)
4.01"
(102)
B
A
C
E
L
K
A0003219
DN
L
A
B
C
K
E
EN (DIN) / JIS /
AS*
[mm]
ANSI
[inch]
inches
(mm)
inches
(mm)
inches
(mm)
inches
(mm)
inches
(mm)
inches
(mm)
15
1/2"
7.87 (200)
11.2 (286)
7.95 (202)
3.31 (84)
4.72 (120)
3.7 (94)
25
1"
7.87 (200)
11.2 (286)
7.95 (202)
3.31 (84)
4.72 (120)
3.7 (94)
32
−
7.87 (200)
11.2 (286)
7.95 (202)
3.31 (84)
4.72 (120)
3.7 (94)
40
1 1/2"
7.87 (200)
11.2 (286)
7.95 (202)
3.31 (84)
4.72 (120)
3.7 (94)
50
2"
7.87 (200)
11.2 (286)
7.95 (202)
3.31 (84)
4.72 (120)
3.7 (94)
65
−
7.87 (200)
13.2 (336)
8.94 (227)
4.29 (109)
7.08 (180)
3.7 (94)
80
3"
7.87 (200)
13.2 (336)
8.94 (227)
4.29 (109)
7.08 (180)
3.7 (94)
100
4"
9.84 (250)
13.2 (336)
8.94 (227)
4.29 (109)
7.08 (180)
3.7 (94)
125
−
9.84 (250)
16.4 (417)
10.5 (267)
5.90 (150)
10.2 (260)
5.51 (140)
150
6"
11.8 (300)
16.4 (417)
10.5 (267)
5.90 (150)
10.2 (260)
5.51 (140)
200
8"
13.8 (350)
18.6 (472)
11.5 (292)
7.08 (180)
12.7 (324)
6.14 (156)
250
10"
17.7 (450)
20.5 (522)
12.5 (317)
8.07 (205)
15.7 (400)
6.14 (156)
300
12"
19.7 (500)
22.5 (572)
13.5 (342)
9.05 (230)
18.1 (460)
6.53 (166)
The fitting length (L) is always the same, regardless of the pressure rating.
* Only DN 25 and 50 are available if flanges according to AS are used.
High-temperature version ≤ 12" (DN ≤ 300)
4.33"
(110)
B1
A1
F06-xxPxxxxx-06-05-00-xx-000
Measurement A1, B1 = Measurement A, B of the standard compact version plus 4.33" (110 mm)
28
Endress+Hauser
Proline Promag 50P, 53P
Compact version ≥ 14" (DN ≥ 350)
8.94"
(227)
8.15" *
(207)
7.68"
(195)
6.61"
(168)
6.30"
(160)
Esc
-
+
E
* Blind version
B
A
C
E
L
K
F06-53Fxxxxx-06-00-xx-xx-001
DN
L
A
B
C
K
E
EN (DIN)
[mm]
ANSI
[inch]
inches
(mm)
inches
(mm)
inches
(mm)
inches
(mm)
inches
(mm)
inches
(mm)
350
14"
21.6 (550)
29.1 (738.5)
17.9 (456.5)
11.1 (282.0)
22.2 (564)
10.9 (276)
400
16"
23.6 (600)
31.1 (790.5)
18.9 (482.5)
12.1 (308.0)
24.2 (616)
10.9 (276)
450
18"
25.6 (650)
33.1 (840.5)
19.9 (507.5)
13.1 (333.0)
26.2 (666)
11.5 (292)
500
20"
25.6 (650)
35.1 (891.5)
20.9 (533.0)
14.1 (358.5)
28.2 (717)
11.5 (292)
600
24"
30.7 (780)
39.2 (995.5)
23.0 (585.0)
16.2 (410.5)
32.3 (821)
15.8 (402)
The fitting length (L) is always the same, regardless of the pressure rating.
Endress+Hauser
29
Proline Promag 50P, 53P
Remote version ≥ 14" (DN ≥ 350)
6.73"
(171)
5.08"
(129)
5.63"
(143)
B
A
C
E
K
L
A0003220
DN
L
A
B
C
K
E
EN (DIN)
[mm]
ANSI
[inch]
inches
(mm)
inches
(mm)
inches
(mm)
inches
(mm)
inches
(mm)
inches
(mm)
350
14"
21.6 (550)
26.9 (683.5)
15.8 (401.5)
11.1 (282.0)
22.2 (564)
10.9 (276)
400
16"
23.6 (600)
29.0 (735.5)
16.8 (427.5)
12.1 (308.0)
24.2 (616)
10.9 (276)
450
18"
25.6 (650)
30.9 (785.5)
17.8 (452.5)
13.1 (333.0)
26.2 (666)
11.5 (292)
500
20"
25.6 (650)
32.9 (836.5)
18.8 (478.0)
14.1 (358.5)
28.2 (717)
11.5 (292)
600
24"
30.7 (780)
37.0 (940.5)
20.9 (530.0)
16.2 (410.5)
32.3 (821)
15.8 (402)
The fitting length (L) is always the same, regardless of the pressure rating.
30
Endress+Hauser
Proline Promag 50P, 53P
Ground disk 1/2" to 12" (DN 15 to 300)
H
ØB
Ø 0.26”
(6.5)
0.59”
(15)
Ø di
0.39”
(10)
ØD
0.08”
(2)
A0003221
DN 1)
di
B
D
H
inches
(mm)]
inches
(mm)
EN (DIN) / JIS /
AS 4)
[mm]
ANSI
[inch]
inches
(mm)
inches
(mm)
15
1/2"
0.63 (16)
1.69 (43)
2.42 (61.5)
2.87 (73)
25
1"
1.02 (26)
2.44 (62)
3.05 (77.5)
3.44 (87.5)
32
−
35
80
87.5
94.5
40
1 1/2"
1.61 (41)
3.23 (82)
3.98 (101)
4.06 (103)
50
2"
2.05 (52)
3.98 (101)
4.55 (115.5)
4.25 (108)
65
−
68
121
131.5
118
80
3"
3.15 (80)
5.16 (131)
6.08 (154.5)
5.31 (135)
100
4"
4.09 (104)
6.14 (156)
7.34 (186.5)
6.02 (153)
125
−
130
187
206.5
160
150
6"
6.22 (158)
8.54 (217)
10.1 (256)
7.24 (184)
200
8"
8.11 (206)
10.5 (267)
11.3 (288)
8.07 (205)
250
10"
300
2)
300 3)
10.2 (260)
12.9 (328)
14.1 (359)
9.45 (240)
2)
12.3 (312)
14.8 (375)
16.3 (413)
10.7 (273)
12" 3)
12.2 (310)
14.8 (375)
15.9 (404)
10.6 (268)
12"
1)
Ground disks can, with the exception of 12" (DN 300), be used for all flange norms / pressure ratings.
PN 10/16, Class 150
3) PN 25, JIS 10K/20K
4) Only DN 25 and 50 are available if flanges according to AS are used.
2)
Endress+Hauser
31
Proline Promag 50P, 53P
Weight
Weight data in kg
Nominal
diameter
Compact version
Remote version (without cable)
Sensor
[mm]
[inch]
EN (DIN) /
AS*
JIS
ANSI
EN (DIN) /
AS*
JIS
Wall
housing
ANSI
6.5
6.5
14 lbs
4.5
4.5
10 lbs
13 lbs
1"
7.3
7.3
16 lbs
5.3
5.3
12 lbs
13 lbs
32
1 1/4"
8.0
7.3
–
6.0
5.3
–
13 lbs
40
1 1/2"
9.4
8.3
20 lbs
7.4
6.3
16 lbs
13 lbs
50
2"
10.6
9.3
23 lbs
8.6
7.3
19 lbs
13 lbs
65
2 1/2"
12.0
80
3"
14.0
100
4"
125
PN 40
1/2"
PN 40
15
25
16.0
14.7
35 lbs
14.0
12.7
31 lbs
13 lbs
5"
21.5
21.0
–
19.5
19.0
–
13 lbs
150
6"
25.5
24.5
200
8"
45
41.9
56 lbs
23.5
22.5
52 lbs
13 lbs
99 lbs
43
39.9
95 lbs
13 lbs
250
10"
65
69.4
165 lbs
63
67.4
161 lbs
13 lbs
300
12"
70
72.3
242 lbs
68
70.3
238 lbs
13 lbs
350
14"
400
16"
115
386 lbs
113
381 lbs
13 lbs
135
452 lbs
133
448 lbs
13 lbs
450
18"
175
562 lbs
173
558 lbs
13 lbs
500
20"
175
628 lbs
173
624 lbs
13 lbs
600
24"
235
893 lbs
233
888 lbs
13 lbs
PN 10
PN 10
Class 150
13 lbs
13 lbs
10K
–
24 lbs
PN 16
9.1
10.5
Class 150
10.0
12.0
10K
–
31 lbs
PN 16
11.1
12.5
Transmitter Promag (compact version): 7.5 lbs (3.4 kg)
High-temperature version: +3.3 lbs (1.5 kg)
(Weight data valid for standard pressure ratings and without packaging material)
* Only DN 25 and 50 are available if flanges according to AS are used.
Materials
Transmitter housing:
• Compact housing: powder coated die-cast aluminium or stainless steel field housing
• Wall-mounted housing: powder coated die-cast aluminium
Sensor housing:
• 1/2" to 12" (DN 15 to 300): powder-coated die-cast aluminium
• 14" to 24" (DN 350 to 600): painted steel (Amerlock 400)
Measuring tube:
• < 14" (DN < 350): stainless steel 1.4301 or 1.4306/304L; non-stainless flange material with
AI/Zn protective coating
• > 12" (DN > 300): stainless steel 1.4301/304; non-stainless flange material with
Amerlock 400 paint
Flange:
• EN 1092-1 (DIN 2501): 316L / 1.4571; RSt37-2 (S235JRG2) / C22 / FE 410W B
(with flanges made of carbon steel: DN < 350 with Al/Zn protective coating,
DN > 300 with Amerlock 400 paint)
• ANSI: A105, F316L SS
with flanges made of carbon steel: < 14" (DN < 350) with Al/Zn protective coating,
> 12" (DN > 300) with Amerlock 400 paint
• JIS: RSt37-2 (S235JRG2) / H II / 1.0425 / 316L
(with flanges made of carbon steel: DN < 350 with Al/Zn protective coating,
DN > 300 with Amerlock 400 paint)
32
Endress+Hauser
Proline Promag 50P, 53P
• AS 2129: (DN 25) A105 or RSt37-2 (S235JRG2)
(DN 50) A105 oder St44-2 (S275JR)
(with Al/Zn protective coating)
• AS 4087: A105 or St44-2 (S275JR)
(with Al/Zn protective coating)
Ground disks: 1.4435/316L SSor Alloy C-22
Electrodes: 1.4435/316L SS, platinum/rhodium 80/20 or Alloy C-22, tantalum
Seals: Seals to DIN EN 1514-1
Material load diagrams
Caution!
The following diagrams contain material load curves (reference curves) for various process connections relating
to the fluid temperature. But the maximal permissible fluid temperature always depends on the lining material
of the sensor and/or the sealing material (s. Page 20).
Flange connection to EN 1092-1 (DIN 2501)
Flange material: RSt37-2 (S235JRG2) / C22 / FE 410W B
[bar]
40
PN 40
35
30
25
PN 25
20
15
PN 16
10
PN 10
5
PN 6
0
-60 -40 -20
0
20
40
60
80 100 120 140 160 180 [°C]
F06-xxFxxxxx-05-xx-xx-xx-000
Flange connection to EN 1092-1 (DIN 2501)
Flange material: 316L / 1.4571
[bar]
40
PN 40
35
30
25
PN 25
20
15
10
5
PN 16
PN 10
PN 6
0
-60 -40
-20
0
20
40
60
80 100 120 140 160 180 [°C]
F06-xxFxxxxx-05-xx-xx-xx-001
Endress+Hauser
33
Proline Promag 50P, 53P
Flange connection to ANSI B16.5
Flange material: A105
[p si]
[bar]
870
60
725
50
Class 300
580
40
435
30
290
20
Class 150
10
145
0
0
-76
-60
-40
-40
32
0
-4
-20
104
40
68
20
140
60
176 212
80 100
248
120
284
140
320 356
160 180
[°F]
[°C]
A0003226
Flange connection to ANSI B16.5
Flange material: F316L SS
[bar]
[psi]
870
60
725
50
580
40
Class 300
30
435
20
290
Class 150
145
10
0
0
-40
-40
-76
-60
-4
-20
32
0
68
20
104
40
140
60
176 212
80 100
248
120
284
140
320
160
356
180
[°F]
[°C]
F06-xxFxxxxx-05-xx-xx-xx-002
Flange connection to JIS B2238
Flange material: RSt37-2 (S235JRG2) / H II / 1.0425
[bar]
30
20K
20
10K
10
0
-60 -40 -20
0
20
40
60
80 100 120 140 160 180 [°C]
A0003228
Flange connection to AS2129 Table E or AS4087 Cl. 14
Flange material: A105 / RSt37-2 (S235JRG2) / St44-2 (S275JR)
[bar]
25
20
15
10
5
0
-60 -40
-20
0
20
40
60
80 100 120 140 160 [°C]
F06-xxFxxxxx-05-xx-xx-xx-010
34
Endress+Hauser
Proline Promag 50P, 53P
Fitted electrodes
Measuring, reference and EPD electrodes
• Standard available with 1.4435/316L SS, Alloy C-22, tantalum, platinum/rhodium 80/20
• Optional: measuring electrodes made of platinum/rhodium 80/20
Process connection
Flange connection:
• EN 1092-1 (DIN 2501), < DN 350 From A, > DN 300 From B
(dimensions acc. to DIN 2501; DN 65 PN 16 and DN 600 PN 16 exclusively to EN 10921)
• ANSI B16.5
• JIS B2238
• AS2129 Table E
• AS4087 Cl. 14
Surface roughness
• PFA liner: ≤ 16 μin (≤ 0.4 μm)
• Electrodes:
– 1.4435/316L SS, Alloy C-22: 12 to 10 μin (0.3 to 0.5 μm)
– Tantalum, platinum/rhodium: 12 to 10 μin (0.3 to 0.5 μm)
(all data relate to parts in contact with the medium)
Human interface
Display elements
• Liquid-crystal display: backlit, two lines (Promag 50) or four lines (Promag 53) with 16 characters per line
• Custom configurations for presenting different measured-value and status variables
• Totalizer:
Promag 50: 2 totalizers (7-digit plus 7-digit overflow with sign and units)
Promag 53: 3 totalizers (7-digit plus 7-digit overflow with sign and units)
Operating elements
Unified operation concept for both types of transmitter:
Promag 50:
• Local operation with three push buttons (–, +, E)
• Quick Setup menus for straightforward commissioning
Promag 53:
• Local operation with Touch Control (–, +, E)
• Application-specific Quick Setup menus for straightforward commissioning
Language group
Language groups available for operation in different countries:
Promag 50, Promag 53:
• Western Europe and America (WEA):
English, German, Spanish, Italian, French, Dutch and Portuguese
• Eastern Europe and Scandinavia (EES):
English, Russian, Polish, Norwegian, Finnish, Swedish and Czech
• South and east Asia (SEA):
English, Japanese, Indonesian
Promag 53:
• China (CIN):
English, Chinese
You can change the language group via the operating program “ToF Tool - Fieldtool Package.”
Remote operation
Endress+Hauser
Promag 50: Remote control via HART, PROFIBUS DP/PA
Promag 53: Remote control via HART, PROFIBUS DP/PA, MODUBUS RS485, FOUNDATION Fieldbus
35
Proline Promag 50P, 53P
Certificates and approvals
Ex approvals
Information about currently available Ex versions (ATEX, FM, CSA) can be supplied by your Endress+Hauser
Sales Centre on request. All explosion protection data are given in a separate documentation which is available
upon request.
CE mark
The measuring system is in conformity with the statutory requirements of the EC Directives. Endress+Hauser
confirms successful testing of the device by affixing to it the CE mark.
C-Tick mark
The measuring system is in conformity with the EMC requirements of the Australian Communications
Authority (ACA).
Pressure Equipment Directive
Flow meters with a nominal diameter smaller or equal DN 25 are covered by Art. 3(3) of the European directive
97/23/EG (Pressure Equipment Directive) and are designed according to sound engineer practice.
For larger nominal diameter, optional approvals according to Cat. III are available when required (depends on
fluid and process pressure).
PROFIBUS DP/PA
certification
The flow device has successfully passed all the test procedures carried out and is certified and registered by the
PNO (PROFIBUS User Organisation). The device thus meets all the requirements of the following specifications:
• Certified to PROFIBUS PA, profile version 3.0 (device certification number: on request)
• The device can also be operated with certified devices of other manufacturers (interoperability)
FOUNDATION Fieldbus
certification
The flow device has successfully passed all the test procedures carried out and is certified and registered by the
Fieldbus Foundation. The device thus meets all the requirements of the following specifications:
•
•
•
•
•
Certified to FOUNDATION Fieldbus Specification
The device meets all the specifications of the FOUNDATION Fieldbus H1.
Interoperability Test Kit (ITK), revision status 4.0 (device certification number: on request)
The device can also be operated with certified devices of other manufacturers
Physical Layer Conformance Test of the Fieldbus Foundation
MODBUS certification
The measuring device meets all the requirements of the MODBUS/TCP conformity test and has the
“MODBUS/TCP Conformance Test Policy, Version 2.0”. The measuring device has successfully passed all the
test procedures carried out and is certified by the “MODBUS/TCP Conformance Test Laboratory” of the
University of Michigan.
Other standards, guidelines
EN 60529:
Degrees of protection by housing (IP code)
EN 61010:
Protection Measures for Electrical Equipment for Measurement, Control, Regulation and
Laboratory Procedures.
EN 61326/A1 (IEC 6326):
Electromagnetic compatibility (EMC requirements)
NAMUR NE 21:
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment.
NAMUR NE 43:
Standardisation of the signal level for the breakdown information of digital transmitters with analogue
output signal.
NAMUR NE 53
Software of field devices and signal-processing devices with digital electronics.
36
Endress+Hauser
Proline Promag 50P, 53P
Accessories
Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter
and the sensor. The Endress+Hauser service organisation can provide detailed information on request.
Supplementary documentation
•
•
•
•
•
•
•
•
Flow Measurement (FA005D/06/en)
Operating Instructions Promag 50 (BA046D/06/en, BA049D/06/en)
Operating Instructions Promag 50 PROFIBUS PA (BA055D/06/en, BA056D/06/en)
Operating Instructions Promag 53 (BA047D/06/en, BA048D/06/en)
Operating Instructions Promag 53 PROFIBUS DP/PA (BA053D/06/en, BA054D/06/en)
Operating Instructions Promag 53 FOUNDATION Fieldbus (BA051D/06/en, BA052D/06/en)
Operating Instructions Promag 53 MODBUS (BA117D/06/en und BA118D/06/en)
Supplementary documentation on Ex-ratings: ATEX, FM, CSA, etc.
Registered trademarks
HART®
Registered trademark of HART Communication Foundation, Austin, USA
PROFIBUS®
Registered trademark of the PROFIBUS User Organisation, Karlsruhe, Germany
FOUNDATION™ Fieldbus
Registered trademark of the Fieldbus FOUNDATION, Austin, USA
MODBUS®
Registered trademark of the MODBUS Organisation
HistoROM™, S-DAT®, T-DAT™, F-CHIP®, ToF Tool - Fieldtool® Package, Fieldcheck®, Applicator®
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH
Endress+Hauser
37
Proline Promag 50P, 53P
Ordering Information
NOTE: Endress+Hauser reserves the right to change or modify product, specifications, and ordering information at any time without notice.
Please consult Endress+Hauser or your local representative for the most recent information.
Promag 50P Electromagnetic flowmeter
0
Promag 50 P
1
2
3
4
5
6
7
8
9
10
11
12
7
Housing
A
NEMA 4X (IP 67) compact aluminum housing
C
NEMA 4X (IP 67) remote wall-mounted (only for approvals A or R)
G
NEMA 4X (IP 67) remote aluminum field housing for non-hazardous areas
K
NEMA 6P sensor, wall-mounted housing (only for approvals A or R)
N
NEMA 6P sensor, aluminum field housing, non-hazardous, not for
liner B high temp. (only up to 12” sensor)
P
NEMA 4X (IP 67) compact aluminum housing, HE (harsh environment),
compact and remote sensor sizes up to 12” only *
S
NEMA 4X (IP 67) remote wall-mounted housing, HE (harsh environment),
remote sensor sizes 14” to 24” (only for approvals A and R),
available as TSP only *
T
NEMA 4X (IP 67) remote field housing, HE (harsh environment),
remote sensor sizes 1/2” to 12” only *
1
-40°F (-40°C) ambient temperature, NEMA 4X (IP 67), compact,
aluminum field housing (only for approvals A and R)
3
-40°F (-40°C) ambient temperature, NEMA 4X (IP 67), compact,
(only for approvals A and R)
5
-40°F (-40°C) ambient temperature, NEMA 6P sensor, wall mount
housing for transmitter (only for approvals A and R)
6
-40°F (-40°C) ambient temperature, NEMA 4X (IP 67) remote field
housing (only for approvals A and R)
Cable for remote
0
Without cable
1
15 foot coil and signal cable
2
30 foot coil and signal cable
5
Coil and signal cable, specify length (maximum 650 ft depending on
conductivity of process material)
Cable entries
B
1/2” NPT
L
1/2” NPT fieldbus connector (only for approval A and R)
X
Sensor only, without cable, compact version, up to 8” only
Power supply / display
7
85 to 260 VAC, without display, remote configuration only (not for
wall mount or SS housing)
8
20 to 55 VAC / 16 to 62 VDC, without display, remote configuration
only (not for wall mount or SS housing)
A
85 to 260 VAC, with display, push button operation (language: EN,
ES, FR, IT, NL, PT, DE)
B
20 to 55 VAC / 16 to 62 VDC, with display, push button operation
(language: EN, ES, FR, IT, NL, PT, DE)
X
Sensor only (without transmitter, only available up to 8”)
Software
A
Standard software
C
ECC electrode cleaning circuit (only for approvals A and R)
Outputs / Inputs
A
Current HART, frequency
D
Current HART, frequency, status output, status input
H
PROFIBUS-PA (approvals A and R only)
J
PROFIBUS-DP (approvals A and R only)
S
Current HART, frequency; active (not for approvals A and R)
T
Current HART, frequency; passive (not for approvals A and R)
W
Current HART
X
Sensor only
-
Nominal Diameter
0
15
1/2”
25
1”
40
1-1/2”
50
2”
80
3”
1H 4”
1F
6”
2H 8”
2F
10”
3H 12”
3F
14”
4H 16”
4F
18”
5H 20”
6H 24”
1
Liner
A
PFA liner, standard (available for 1” to 8” sensors)
B
PFA liner, high temperature version (available for 1” to 8” sensors,
not for approval N)
E
PTFE liner
2
Process connection
L
Class 150 ANSI B16.5 CS steel A105 flanges
M
Class 300 ANSI B16.5 CS steel A105 flanges (not for 8” and larger)
R
Class 150 ANSI B16.5 316L SS flanges
S
Class 300 ANSI B16.5 316L SS flanges (not for 8” and larger)
3
Electrodes / material
0
Measuring, reference and EPD electrodes / 316L SS
1
Measuring, reference and EPD electrodes / Alloy C22
2
Measuring, reference and EPD electrodes / tantalum
3
Measuring electrodes only, Platinum/Rhodium 80/20% (not available for
14” to 24”)
5
Measuring, reference and EPD electrodes, Platinum/Rhodium 80/20% (not
available for 14” to 24”)
G
Measuring, reference and EPD electrodes, bullet nose / 316L SS
H
Measuring, reference and EPD elecctrodes, bullet nose / Alloy C-22
4
Calibration
A
3-point calibration, 0.5%
B
3-point calibration, 0.2%
D
SCS/A2LA 3-point 0.5% calibration (ISO/IEC 17025) with certificate
traceable according to ISO 9000 (specify range)
E
SCS/A2LA 3-point 0.2% calibration (ISO/IEC 17025) with certificate
traceable according to ISO 9000 (specify range)
5
Certificates
1
Standard, no certificate
2
3.1B material certificate for pipe and flanges
3
2.3 pressure test certificate (1.5 x PN, 3 minutes)
4
3.1B material and 2.3 pressure test certificate
6
Approvals
A
For use in non-hazardous areas
N
FM explosion proof Class I, Div. 1 / CSA Class I, Div. 1 (only for aluminum
field housing, compact version, not for 14” and larger sensors)
R
FM non-incendive Class I, Div. 2 / CSA Class I, Div. 2
8
9
10
11
12
*
38
Harsh environment (HE) option is available for process conditions where
cool process temperatures in tropical (high humidity) environments or process
fluids which undergo large cyclical temperature variations which can cause
high amounts of moisture that could condense onto the measurement tube.
Endress+Hauser
Proline Promag 50P, 53P
Promag 53P Electromagnetic flowmeter
0
Promag 53 P
1
2
3
4
5
6
7
8
9
10
11
12
7
Housing
A
NEMA 4X (IP 67) compact aluminum housing
C
NEMA 4X (IP 67) remote wall-mounted (only for approvals A or R)
G
NEMA 4X (IP 67) remote aluminum field housing for non-hazardous areas
K
NEMA 6P sensor, wall-mounted housing (only for approvals A or R)
N
NEMA 6P sensor, aluminum field housing, non-hazardous, not for
liner B high temp. (only up to 12” sensor)
P
NEMA 4X (IP 67) compact aluminum housing, HE (harsh environment),
compact and remote sensor sizes up to 12” only *
S
NEMA 4X (IP 67) remote wall-mounted housing, HE (harsh environment),
remote sensor sizes 14” to 24” (only for approvals A and R),
available as TSP only *
T
NEMA 4X (IP 67) remote field housing, HE (harsh environment),
remote sensor sizes 1/2” to 12” only *
1
-40°F (-40°C) ambient temperature, NEMA 4X (IP 67), compact,
aluminum field housing (only for approvals A and R)
3
-40°F (-40°C) ambient temperature, NEMA 4X (IP 67), compact,
(only for approvals A and R)
5
-40°F (-40°C) ambient temperature, NEMA 6P sensor, wall mount
housing for transmitter (only for approvals A and R)
6
-40°F (-40°C) ambient temperature, NEMA 4X (IP 67) remote field
housing (only for approvals A and R)
Cable for remote
0
Without cable
1
15 foot coil and signal cable
2
30 foot coil and signal cable
5
Coil and signal cable, specify length (maximum 650 ft depending on
conductivity of process material)
Cable entries
B
1/2” NPT
L
1/2” NPT fieldbus connector (only for approval A and R)
X
Sensor only, without cable, compact version, up to 8” only
Power supply / display
7
85 to 260 VAC, without display, remote configuration only (not for
wall mount or SS housing)
8
20 to 55 VAC / 16 to 62 VDC, without display, remote configuration
only (not for wall mount or SS housing)
A
85 to 260 VAC, with display, push button operation (language: EN,
ES, FR, IT, NL, PT, DE)
B
20 to 55 VAC / 16 to 62 VDC, with display, push button operation
(language: EN, ES, FR, IT, NL, PT, DE)
X
Sensor only (without transmitter, only available up to 8”)
Software
A
Standard software
C
ECC electrode cleaning circuit (only for approvals A and R)
Outputs / Inputs
Fixed communication boards
A
Current HART, frequency
B
Current HART, frequency, 2 relays
F
PROFIBUS-PA, IS
G
Foundation Fieldbus, IS
H
PROFIBUS-PA
J
PROFIBUS-DP
K
Foundation Fieldbus
Q
Modbus RS485, status input
S
Current HART, frequency; active I.S
T
Current HART, frequency; passive I.S
Flexible communication boards
C
Current HART, frequency, 2 relays, flexible module
D
Current HART, frequency, relay, and status input, flexible module
L
Current HART, 2 relays and status input/output
M
Current HART, 2 frequency output, status input
N
Modbus RS 485, current and frequency output, status input
P
PROFIBUS-DP, current and frequency output, status input
V
PROFIBUS-DP, two relay outputs, status input
2
Current HART, relay, current, frequency ouputs
4
Current HART, relay, frequency outputs, current input
7
Modbus RS485, two relay outputs, status input
X
Sensor only
-
Nominal Diameter
0
15
1/2”
2F 10”
25
1”
3H 12”
40
1-1/2”
3F 14”
50
2”
4H 16”
80
3”
4F 18”
1H 4”
5H 20”
1F
6”
6H 24”
2H 8”
1
Liner
A
PFA liner, standard (available for 1” to 8” sensors)
B
PFA liner, high temperature version (available for 1” to 8” sensors,
not for approval N)
E
PTFE liner
2
Process connection
L
Class 150 ANSI B16.5 CS steel A105 flanges
M
Class 300 ANSI B16.5 CS steel A105 flanges (not for 8” and larger)
R
Class 150 ANSI B16.5 316L SS flanges
S
Class 300 ANSI B16.5 316L SS flanges (not for 8” and larger)
3
Electrodes / material
0
Measuring, reference and EPD electrodes / 316L SS
1
Measuring, reference and EPD electrodes / Alloy C22
2
Measuring, reference and EPD electrodes / tantalum
3
Measuring electrodes only, Platinum/Rhodium 80/20% (not available for
14” to 24”)
5
Measuring, reference and EPD electrodes, Platinum/Rhodium 80/20% (not
available for 14” to 24”)
G
Measuring, reference and EPD electrodes, bullet nose / 316L SS
H
Measuring, reference and EPD elecctrodes, bullet nose / Alloy C-22
4
Calibration
B
3-point calibration, 0.2%
E
SCS/A2LA 3-point 0.2% calibration (ISO/IEC 17025) with certificate
traceable according to ISO 9000 (specify range)
5
Certificates
1
Standard, no certificate
2
3.1B material certificate for pipe and flanges
3
2.3 pressure test certificate (1.5 x PN, 3 minutes)
4
3.1B material and 2.3 pressure test certificate
6
Approvals
A
For use in non-hazardous areas
N
FM explosion proof Class I, Div. 1 / CSA Class I, Div. 1 (only for aluminum
field housing, compact version, not for 14” and larger sensors)
R
FM non-incendive Class I, Div. 2 / CSA Class I, Div. 2
8
9
10
11
12
*
Housing options
Harsh environment (HE) option is available for process conditions where
cool process temperatures in tropical (high humidity) environments or process
fluids which undergo large cyclical temperature variations which can cause
high amounts of moisture that could condense onto the measurement tube.
Endress+Hauser
39
Subject to modification
International Head Quarter
Endress+Hauser
GmbH+Co. KG
Instruments International
Colmarer Str. 6
79576 Weil am Rhein
Deutschland
United
States
Tel. +49
76 21 9 75 02
Fax +49 76 21 9 75 34 5
Endress+Hauser, Inc.
www.endress.com
2350 Endress Place
[email protected]
Greenwood,
IN 46143
Tel. 317-535-7138
Sales 888-ENDRESS
Service 800-642-8737
fax 317-535-8498
[email protected]
www.us.endress.com
TI047D/24/ae/07.08
© 2008 Endress+Hauser
Canada
Mexico
Endress+Hauser Canada
1075 Sutton Drive
Burlington, ON L7L 5Z8
Tel. 905-681-9292
800-668-3199
Fax 905-681-9444
www.ca.endress.com
Endress+Hauser, México, S.A. de C.V.
Fermando Montes de Oca 21 Edificio A Piso 3
Fracc. Industrial San Nicolás
54030. Tlalnepantla de Baz
Estado de México
México
Tel: +52 55 5321 2080
Fax +52 55 5321 2099
[email protected]
www.mx.endress