technical manual

Add to My manuals
134 Pages

advertisement

technical manual | Manualzz
UV FLUORESCENT
PRINCIPLE OF
OPERATION
GENERAL
INFORMATION
CARACTERISTICS
AF22M
OPERATING
INSTRUCTION
TECHNICAL MANUAL
APPENDIX
CORRECTIVE
MAINTENANCE
- JUNE 2010 -
PREVENTIVE
MAINTENANCE
SULFUR DIOXIDE ANALYZER
111 bd Robespierre, 78300 POISSY - -TEL. 33(0)-1.39.22.38.00 – FAX 33(0)-1.39 65.38.08
http://www.environnement-sa.com
AF22 MODULE
Environnement S.A
Duplication prohibited
0
WARNING
The information in this documentation is subject to change without notice.
Environnement S.A. all rights reserved.
This document does not represent a commitment under part of Environnement S.A.
0–2
JUNE 2004
AF22 MODULE
Duplication prohibited
Environnement S.A
SUMMARY
CHAPTER 1 GENERAL INFORMATION - CHARACTERISTICS
1.1.
GENERAL INFORMATION
1.2.
CHARACTERISTICS
13
110
CHAPTER 2 PRINCIPLE OF OPERATION
2.1.
THEORETICAL BASIS
23
2.2.
PRINCIPLE OF MEASUREMENT
25
2.3.
TAKING SAMPLES AND ANALYSIS
27
2.4.
SIMPLIFIED FLOW CHART OF PRINCIPAL PROGRAM
29
2.5.
AUTOMATIC RESPONSE TIME
210
CHAPTER 3 OPERATING INSTRUCTIONS
3.1.
INITIAL START-UP
34
3.2.
PROGRAMMING THE AF22M
37
3.3.
DESCRIPTION OF THE DIFFERENT SCREENS
310
3.4.
CALIBRATION
341
CHAPITRE 4 PREVENTIVE MAINTENANCE
4.1.
SAFETY INSTRUCTIONS
42
4.2.
MAINTENANCE CALENDAR
43
4.3.
MAINTENANCE OPERATION SHEETS
44
4.4.
AF22M MAINTENANCE KIT
411
CHAPTER 5 CORRECTIVE MAINTENANCE
CHAPTER 6 APPENDIX
ESTEL BOARD
SOREL BOARD
JUNE 2004
0–3
AF22 MODULE
Environnement S.A
Duplication prohibited
LIST OF TABLES
Table 3.1 – DB37 and DB25 connectors links
Table 3.2 – MUX signals (Acceptable limit on the multiplexer 1 to 16 channels)
Table 5–1– List of faults and corrective actions
Table 5–2 – AF22M Module board configuration
Table 5–3 – Board RS4i Configuration
Table 5–4 – Keyboard Interface configuration
Table 5–5 – UV lamp board configuration
Table 5–6 – Configuration of UV lamp supply board
Table 5–7 – Flow rate control board configuration and setting
33
337
5–4
5–9
5–10
5–11
5–12
5–13
5–14
LIST OF FIGURES
Figure 1.1 – AF22M Presentation
Figure 1.2 – Keyboard and display
Figure 1.3 – Rear panel
Figure 1.4 – Components location
Figure 1.5 – Links between units
Figure 1.6 – Outline dimensions
Figure 2.1 – Diagram showing molecule energy levels
Figure 2.2 – General principle diagram
Figure 2.3 – Filtration of hydrocarbon molecules
Figure 3.1 – Electrical connections
Figure 3.2 – Fluid connection
Figure 3.3 – Permeation tube installation
Figure 3.4 – Software overview
Figure 3.5 – Printout example
Figure 3.6 – Pressurized gas connection example
Figure 3.7 – Typical calibrator
Figure 4.1 – Inlet dust filter
Figure 4.2 – Zero filter cartridges position
Figure 4.3 – Exploded view of the pump
Figure 4.4 – Lamp/shutter assembly (front face)
Figure 4.5 – Lamp/shutter assembly (rear face)
Figure 5-1 – AF22M Module board
Figure 5-2 – Card RS4i Configuration
Figure 5-3 – Keyboard Interface board
Figure 5-4 – UV lamp board
Figure 5–5 – UV lamp supply board
Figure 5-6 – Flow rate control board
0–4
12
13
14
15
19
110
22
26
27
33
3–4
3–5
3–9
3–34
3–42
3–45
4–4
4–7
4–9
4–10
4–10
5–8
5–10
5–11
5–12
5–13
5–14
MAY 2007
AF22 MODULE
Duplication prohibited
Environnement S.A
INDEX OF PAGES
Page
Date
Page
Date
0-1
0-2
0-3
0-4
0-5
0-6
06.2010
07.2002
06.2004
05.2007
06.2010
06.2010
1-1
1-2
1-3
1-4
1-5
1-6
1-7
1-8
05.2007
07.2002
07.2002
05.2007
07.2002
05.2007
05.2007
07.2002
2-1
2-2
2-3
2-4
2-5
2-6
2-7
2-8
2-9
2-10
05.2007
07.2002
07.2002
07.2002
05.2007
07.2002
07.2002
07.2002
07.2002
07.2002
3-27
3-28
3-29
3-30
3-31
3-32
3-33
3-34
3-35
3-36
3-37
3-38
3-39
3-40
3-41
3-42
3-43
3-44
3-45
3-46
3-47
3-48
3-49
3-50
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
05.2007
01.2008
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
3-1
3-2
3-3
3-4
3-5
3-6
3-7
3-8
3-9
3-10
3-11
3-12
3-13
3-14
3-15
3-16
3-17
3-18
3-19
3-20
3-21
3-22
3-23
3-24
3-25
3-26
07.2002
05.2007
07.2002
05.2007
07.2002
05.2007
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
05.2007
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
05.2007
05.2007
07.2002
05.2007
07.2002
4-1
4-2
4-3
4-4
4-5
4-6
4-7
4-8
4-9
4-10
4-11
4-12
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
5-1
5-2
5-3
5-4
5-5
5-6
5-7
5-8
5-9
5-10
5-11
5-12
5-13
5-14
05.2007
07.2002
07.2002
07.2002
05.2007
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2002
07.2007
06.2004
JUNE 2010
Page
Date
6-1
6-2
06.2004
05.2007
0–5
AF22M
Environnement S.A
Duplication prohibited
EPA EQUIVALENCY DESIGNATION
Environnement S.A. Model AF22M SO2 Analyzer Automated Equivalent Method: EQSA-0802-149
“Environnement S.A Model AF22M UV Fluorescence Sulfur Dioxide Analyzer,” operated with a full
scale range of 0 - 500 ppb, at any temperature in the range of 10 °C to 35 °C, with a 5-micron PTFE
sample particulate filter, with a response time setting of 11 (Automatic response time), with the automatic “ZERO-REF” cycle ON and set for activation every 24 hours, and with or without either of the
following options: Permeation oven, Rack mount / slides.
[Federal Register: Vol. 67, page 57811, 09/12/02]
0–6
JUNE 2010
AF22 MODULE
Duplication prohibited
Environnement S.A
CHAPTER 1
GENERAL INFORMATION - CHARACTERISTICS
1.1
GENERAL INFORMATION
1–3
1.1.1
PRESENTATION
1–3
1.1.2
DESCRIPTION
1–3
1.1.2.1
Front panel
1–3
1.1.2.2
Rear panel
1–4
1.1.2.3
Components locations
1–7
1.1.3
1.1.4
1.2
OPERATING MODES
1–8
1.1.3.1
Standard
1–8
1.1.3.2
Optional
1–8
ASSOCIATED EQUIPMENT
1–8
CHARACTERISTICS
1–9
1.2.1
TECHNICAL CHARACTERISTICS
1–9
1.2.2
OPERATING CHARACTERISTICS
1–10
1.2.3
STORAGE CHARACTERISTICS
1–10
1.2.4
INSTALLATION CHARACTERISTICS
1–10
1.2.4.1
Links between units
1–10
1.2.4.2
Dimensions and weight
1–10
1.2.4.3
Handling and storage
1–10
Figure 1-1 - AF22M Presentation
Figure 1-2 - Keyboard and display
Figure 1-3 – Rear panel
Figure 1-4 - Components location
Figure 1-5 - Links between units
Figure 1-6 – Outline dimensions
MAY 2007
1–2
1–3
1–4
1–6
1–10
1–11
1–1
AF22 MODULE
Environnement S.A
1
Duplication prohibited
GENERAL INFORMATION - CHARACTERISTICS
Figure 1-1 - AF22M Presentation
1–2
JULY 2002
Duplication prohibited
AF22 MODULE
Environnement S.A
1.1
GENERAL INFORMATION
1.1.1
PRESENTATION
The AF22M is a continuous sulfur dioxide monitor designed for use at low contents in ambient air.
The detection principle is based on fluorescence in ultraviolet.
The monitor provides many advantages by the use of recent advanced electronic and optical
technologies and requires very limited maintenance.
The sample is taken with a Teflon tube (outside diameter 6 mm) connected to the back of the unit.
The measurement is indicated by a graphic display on the front panel.
1.1.2
DESCRIPTION
1.1.2.1
Front panel
The front panel includes:
a general switch
a backlit liquid crystal display
– 16 lines 40 columns (240 x 128 pixels)
– the display provides the measurement values according to the selected unit, the information
required for programming and testing the unit.
a keyboard with 6 touch-sensitive keys
The control and check functions of the unit are controlled through the keyboard.
– the function of each key varies with the different screens or menus.
Figure 1-2 - Keyboard and display
JULY 2002
1–3
AF22 MODULE
Environnement S.A
1.1.2.2
Duplication prohibited
Rear panel
The rear panel of the AF22M contains the electrical connectors and gas inlets/outlets.
Gas inlets/outlets (right hand-side)
– The inlet of sample to be analyzed is composed of a PVDF 4/6 mm fitting associated with a dust
filter holder equipped with Teflon filtering membrane (1).
– The "span gas" inlet (4) consists of a 4/6 mm PVDF fitting to connect an outside span gas delivered
at atmospheric pressure, or span gas excess pipe when optional permeation bench is available.
In case of permeation bench option, this inlet becomes an outlet of the permeation
bench (certification de calibration certification by built-in permeation bench).
– The "pump outlet" (3) to exhaust the analyzed sample consists of a 4/6 mm PVDF fitting.
– The optional "zero air inlet" (2) allows to connect an external zero air generator.
NOTE : This inlet is sealed when optional built-in zero filter is present.
Optional : - optional permeation bench zero air inlet (10).
Electrical Equipment Connections (left side)
– The main power supply assembly consists of a 3-contact socket (5) for standard power cable
connection, and general fuse: 0,6 A / 220 V or 1,2 A / 115 V (6)
– 1 standard 25-pin plug (7), for serial links COM1 (RS 232C – RS 422) and COM 2 (RS232C).
– 1 standard 37 pin (8) (see table 3-1 – DB37 and DB connectors links)
Ventilation
Ventilation is achieved by a fan (9).
Figure 1-3 – Rear panel
1–4
MAY 2007
Duplication prohibited
AF22 MODULE
Environnement S.A
Page intentionally blank
JULY 2002
1–5
AF22 MODULE
Environnement S.A
6
3
Duplication prohibited
2
1
17
18
19
20
7
A
20
10
14
16
15
13
7
12
11
Figure 1-4 - Components location
1–6
MAY 2007
AF22 MODULE
Duplication prohibited
1.1.2.3
Environnement S.A
Components locations
The components inside the unit are accessed by simply unscrewing the screws at rear and lateral
sides of the unit and removing the upper cover.
Mechanical components
This includes the following equipment:
– the solenoid valves and filter assembly (1 and 2),
– the internal "Zero Air" filter (activated charcoal) (6),
– the carbon kicker filter (activated charcoal) (18),
– the temperature controlled compartment (17),
– the photo multiplier tube (15),
– the UV lamp (13).
After passing through the inlet dust filter (1), the sample to be analyzed is sent to a block of two 3-way
solenoid valves (2).
These solenoid valves are used to select:
– either in "MEASUREMENT" mode, the gas sampling to be analyzed,
– or in "ZERO" mode, the filtered ambient air through the activated charcoal filter (6),
– or in "SPAN" mode, a span gas originating from:
x the permeation oven (19) if there is one (optional),
The permeation oven consists of an aluminum chamber inside which the permeation tube is
placed. This chamber is regulated at a temperature of 50 °C and ventilated by an additional
constant flow rate pump (20).
x or an outside calibration device.
The sample to be analyzed is drawn through a carbon kicker (14) aromatic hydrocarbon elimination
device to the reaction chamber (16) in which the fluorescence takes place. Fluorescence is detected
by a photo multiplier tube (15).
The temperature of the optical block is regulated at 43 °C by the microprocessor associated with a
probe and a heating resistance.
Outside the compartment, the "zinc ray" type lamp (13) generating the UV radiation is powered by a
stabilized voltage supply (11). The UV beam emitted is interrupted during the beginning of "ZERO
REF" by a shutter (12) in order to measure the PM tube black current.
The analyzed sample passes through a flow regulating restrictor and through the external tube of the
carbon kicker to eliminate hydrocarbon molecules. It is then expelled by a membrane pump (3).
The supply bloc 24V DC (20) enables to supply the whole necessary voltages.
If the user wants to modify the original supply voltage (230 V AC to 155 V AC), 2
options are necessary :
x
switch the A switch to the cut-out power supply
x
modify value of general fuse :
– from 0,6 A passes to 1,2 A for a normal analyzer,
– from 1,6 A passes to 3,2 A if the CH2S option is available.
MAY 2007
1–7
AF22 MODULE
Environnement S.A
Duplication prohibited
Electronic part
– The signal output by the photo multiplier tube is amplified and converted into a digital signal on the
Module board (7). This board also:
x supplies DC voltages at + 15 V, 15 V, + 5 V,
x detects any alarms, transmits these alarms to the microprocessor board and signals them on
the corresponding output connector,
x performs digital/analog conversion of measurement values output by the microprocessor board.
– The microprocessor (7), calculates and stores measurement values, and manages alarms and
automatic cycles.
– The keyboard interface board (10) controls the dialogue between the microprocessor board and the
keyboard and the display unit.
– The UV lamp power supply (11) provides the stabilized voltage necessary for the zinc ray lamp.
The current in the zinc ray lamp is regulated.
1.1.3
OPERATING MODES
1.1.3.1
Standard
– Programmable measurement range up to 10 ppm, with a minimum detectable limit of 1 ppb (typical
value is below 0,5 ppb).
– Remote-controlled or programmable automatic calibration and zero sequence.
– Automatic control of parameters influencing metrology and tests of correct functioning.
3
– Measurement values indicated in ppb or μ/m .
– Memorization of average measurements with a programmable period (capacity 5700 averages).
1.1.3.2
Optional
The monitor can be equipped with the following options:
– Permeation oven.
– ESTEL board(s) :
1.1.4
x
Analog outputs of the SO2 concentration and two of the 16 multiplexer channels.
x
Remote signaling of "measurement", "zero", "calibration" and "alarm" functions.
ASSOCIATED EQUIPMENT
– Analog recorders and data loggers.
– Numerical data acquisition system
– Serial printer for continuous printout of display measurements (programmable period) and
configuration.
1–8
JULY 2002
Duplication prohibited
AF22 MODULE
1.2
CHARACTERISTICS
1.2.1
TECHNICAL CHARACTERISTICS
Environnement S.A
Measurement range (programmable)
:
user programmable up to 10.00 ppm *
Units
:
ppm or mg/m3 (programmable)
Noise (V)
:
0.0005 ppm (RT 120)
Minimum detectable limit (2V)
:
0.001 ppm (RT 120)
Response time (0-90 %)
:
20-120" fixed or automatic (programmable)
Zero drift
:
< 1 ppb./24 hours in operation
Span drift
:
< 1% / 24 hours
Linearity
:
± 1%
H20 influence
:
Null
Temperature influence
:
0.3 ppb/°C
Sample flow rate
:
about 415 cc/min (internal pump)
Display
:
LCD 240x128 text and graphic modes
Control keyboard
:
6 context dependent keys
Output signals
:
3 analog outputs 0-1V, 0-10V, 0-20 mA or 4-20
mA
Power supply
:
230V-50Hz (115V-60Hz) + ground
Consumption
:
280 VA starting up, 110 VA normal operation
Working temperature
:
+ 10 °C to + 35 °C
Memorization of measurement values
:
Capacity: 5700 last averages of the 3 displayed
parameters
Measurement values printing
: By serial printer connected to COM2
Alarms checks
:
– Permanent
– Detection and indication of
functioning anomalies: optical
block temperature, sample flow
rate, UV energy, PM tube high
voltage, SO2 measurement overthreshold, over-range, calibration
fault ...
Tests and diagnostics for maintenance
: Selection at keyboard and display of all
parameters.
Backup saving time for data memorized in
: > 6 months by incorporated battery.
RAM and of the real-time clock.
* 0.500 ppm is the only range covered under EPA equivalency designation.
JULY 2002
1–9
AF22 MODULE
Environnement S.A
1.2.2
OPERATING CHARACTERISTICS
Not applicable.
1.2.3
STORAGE CHARACTERISTICS
Duplication prohibited
– Temperature: 10 ° to 60 °C.
Links between units
The AF22M monitor uses the external links and power supplies illustrated below:
Power
supplies
220V/1A
or
115V/2A
"Sample gas"
1.2.4.1
"Zero air"
(Option)
INSTALLATION CHARACTERISTICS
"Span gas"
1.2.4
AF22M
Serial
link
ESTEL
Interface
(option)
Exhaust or
pump outlet
Figure 1-5 - Links between units
1.2.4.2
Dimensions and weight
The analyzer is a standard 19-inch, 3-unit rack high.
Length
: 591 mm
Width
: 483 mm
Height
: 133 mm
Weight
: 9 kg
1.2.4.3
Handling and storage
The AF22M monitor must be handled with care to avoid damage to the various connectors and fittings
on the rear panel.
Ensure the fluid inlets and outlets on the unit are protected with caps whenever storing the monitor.
The unit is stored in a foam-packed case provided for this purpose.
1–10
JULY 2002
AF22 MODULE
Duplication prohibited
Environnement S.A
430
50
125
5
54
133
14
38
483
57
7
36
Figure 1-6 – Outline dimensions
JULY 2002
1–11
AF22 MODULE
Environnement S.A
Duplication prohibited
Page intentionally blank
1–12
JULY 2002
Duplication prohibited
AF22 MODULE
Environnement S.A
CHAPTER 2
PRINCIPLE OF OPERATION
2.1
THEORETICAL BASIS (Fig. 2.1)
2–3
2.2
PRINCIPLE OF MEASUREMENT
2–5
2.3
TAKING SAMPLES AND ANALYSIS
2–7
2.4
SIMPLIFIED FLOW CHART OF PRINCIPAL PROGRAM
2–9
2.5
AUTOMATIC RESPONSE TIME
2–10
2.5.1
SIMPLIFIED PRINCIPLE OF OPERATION
2–10
2.5.2
PROGRAMMING THE RESPONSE TIME
2–10
Figure 2-1 - Diagram showing molecule energy levels
Figure 2-2 – General principle diagram
Figure 2-3 – Filtration of hydrocarbon molecules
MAY 2007
2–2
2–6
2–7
2–1
AF22 MODULE
Environnement S.A
2
Duplication prohibited
PRINCIPLE OF OPERATION
Electronical
Vibrational
Rotational
Energy levels
A: Electronical spectrum (system of bands associated with an electronic transition)
B: Vibration - rotation spectrum (band associated with a vibrational transition)
C: Rotation spectrum (rays associated with a rotational transition)
Figure 2-1 - Diagram showing molecule energy levels
2–2
JULY 2002
Duplication prohibited
2.1
AF22 MODULE
Environnement S.A
THEORETICAL BASIS (FIG. 2.1)
Energy diagram:
Many developments in quantum mechanics at the beginning of this century enabled theoretical
physicists to conceptualize the processes involved in energy exchanges between a gaseous molecule
such as sulfur dioxide and its environment.
The diagram shown makes it easier to obtain a global understanding of the various phenomena due to
absorption of a radiation by the molecule.
Two essential comments clarify this diagram:
Energy levels are quantified and distributed in accordance with a structure that is different for molecule
electronic, vibrational and rotational levels,
At the considered wavelengths, the "electronic" scale of energies only includes the energy levels of
molecule valency electrons.
Also, a transition from one electronic level to another by absorption of a photon is always
accompanied by lower level vibrational and rotational transitions.
The mathematical formulation of molecular physics associates a spatial and temporal function called a
wave function to each state of the molecule, fully characterizing this state. A transition then occurs by
a new spatial distribution of the wave function.
Absorption and emission:
In its fundamental state Ee0, the SO2 molecule can only absorb sufficient energy photons to access
the first excited state on the electronic scale Ee1.
The energy of a photon is given by Einstein's relation:
E
hv
h
c
O
where O is the radiation wavelength output from a low-pressure zinc vapor lamp, namely O = 213.9 nm,
and h and c represent Plank's constant and the speed of light in a vacuum respectively.
The transition is denoted in the following form:
SO2 + h v o SO2*
As we have already said, the molecule always reaches a vibrational and rotational sub-level of Ee1,
namely higher than this sub-level. It then very quickly dissipates its vibration and rotation energy to
remain on level Ee1 a little longer (a few nanoseconds).
JULY 2002
2–3
AF22 MODULE
Environnement S.A
Duplication prohibited
Starting from Ee1, it can reach any sub-level of its fundamental state along different paths:
by fluorescence:
Kf
SO2 * o SO2 hv'
Based on the above,
firstly v' =
c
O'
is less than v, therefore O' > O,
secondly, O' can be equal to one of several values around an average length, since the sub-level
reached can be arbitrary (statistical distribution as a function of the temperature).
The monitor "observes" photons emitted through a filter centered on 350 nm.
by extinction:
Kq
SO2 * M  o SO2 M
where M represents another gaseous molecule.
The molecule then dissipates its energy mechanically and does not fluoresce. This is referred to as
"quenching".
by dissociation:
SO2 * Kd
o SO O
The energy Ee1 is sufficient to break the SO O link.
Kf, Kd, Kq denote quantum yields associated with each form of deactivation. They are related to
the life of state Ee1, and express probabilities attached to each reaction type. They are determined
from the integral, taken over the entire volume, of the product of wave functions of excited and
fundamental states.
2–4
JULY 2002
Duplication prohibited
2.2
AF22 MODULE
Environnement S.A
PRINCIPLE OF MEASUREMENT
The radiation intensity absorbed by sulfur dioxide in an optical chamber of length L follows the BeerLambert law:
ia = i0 x (1 – e-Lc)
where lo denotes the intensity at the chamber inlet, D is the characteristic SO2 absorption coefficient
and c = [SO2], is the concentration of the gas to be analyzed.
Also, the probability that an excited molecule will fluoresce is given by:
Kf
Kf Kq Kd
The fluorescence intensity received by the PM is then expressed in the following form:
If
G la
Kf
Kf Kq Kd
where G is a constant that depends on the illuminated part of the chamber seen by the PM.
Therefore:
if
In our case, D L c << 1 and 1 e
Gio
DLc
Kf
x (1 e D Lc )
Kf Kq Kd
can be developed to the first order as follows:
1e
DLc
#+DLc
We then obtain:
If
G lo Kf D L
c
Kf Kq Kd
E .c
The radiation captured by the PM is thus directly proportional to the SO2 concentration. This result is
the basis for the measurement technique used with the AF22M monitor.
MAY 2007
2–5
AF22 MODULE
Environnement S.A
Duplication prohibited
Sample inlet
Span gas inlet
(or span gas exess for optional permeation bench)
Optional external zero air inlet
(Sealed for use of built in zero filter)
Exhaust
Dust filter
Activated charcoal filter
Pump
Solenoid valves assembly
~ 40 l/h
Permeation bench
Restrictor
Pump
Aromatic hydrocarbons scrubber
~ 45 l/h
Activated charcoal filter
Flow rate
sensor
Flow rate
sensor
Activated charcoal filter
Built in optional permeation bench assembly
~25 l/h
Black
Stabilized power supply
SO2
Reaction chamber
UV
UV
UV
UV
UV
UV
UV
UV
UV
UV
UV
UV
UV
UV
UV
UV
UV
UV UVUV
UV
UV
UV
UV
UV
UV UV UV
UV
UV UV
UV UV
UV UV UV
UV UV
UV UV UV
UV UV
UV UV
UV UV
UV UVUV UV
UV UVUV UV
UV UV UV UVUV UV UV UV
UV UV
UV
UV UV
UVUVUV
UVUV UV UV
UVUV
UVUVUV
UV UV
UVUVUV UV
UV
UVUVUV UV
UV UV
UV
UV
UV
UV
UV
UV UV
UV
UV
UV
UV
UV
UV
UV
UV
UV
UV
UV
UV
UV
Photomultiplier tube
UV UV
UV
UV
UV
UV UV UV UV
UV UVUV
UV
UV UV
UVUV UV UVUV UVUV UV UV
UV
UV
UV UV UV UV
UV
UV
UV
UV
UV
detector
Modul board
Shutter
UV
UV
UV UV
UV UVUV UV
UV UVUV UV
UVUV
UVUV UV UV
UV UV
UV
UV
UV UV
Zinc ray UV lamp
Restrictor
SO2 Measurements
Figure 2-2 – General principle diagram
2–6
JULY 2002
AF22 MODULE
Duplication prohibited
2.3
Environnement S.A
TAKING SAMPLES AND ANALYSIS
The sample is taken by a Teflon tube connected to the back of the monitor through a pump placed at
the end of the circuit. A Teflon filter provides dust protection.
Filtration of hydrocarbon molecules
The sample to be analyzed is firstly filtered by an elimination device of aromatic hydrocarbon
molecule.
CARBON KICKER
HC
SAMPLE
INLET
HC HC
HC
HC
HC
HC
HC
CHAMBER
HC
HC HC
HC
HC
HC
HC
HC
Activated charcoal filter
Restrictors
Pump
Figure 2-3 – Filtration of hydrocarbon molecules
This device consists of two concentric tubes. The internal tube is made of a special polymer (silicone).
The sample to be analyzed with aromatic HC molecules comes in the internal tube. Aromatic HC
molecules are transferred by permeation to the external tube (silicone) with effect that the transfer is
done in the direction: more HC molecules in gas to few HC molecules in gas.
The pump creates a vacuum in the external tube, the partial pressure of aromatic compounds
decreases due to activated charcoal filtration and molecules are evacuated to the outside of the
internal tube.
JULY 2002
2–7
AF22 MODULE
Environnement S.A
Duplication prohibited
Analysis
The sample to be analyzed, exempt from HC molecules, is directed to a reaction chamber in which it is
irradiated by an ultraviolet radiation centered at 214 nm, the absorption wavelength of SO2 molecules.
A photodiode measures the ultraviolet radiation generated by the UV lamp, through a mirror. This
measurement is used during signal processing in order to compensate for any variation of the UV
energy.
Molecules restore a specific fluorescence in the ultraviolet, which is optically filtered between 300 and
400 nm at the outlet in order to eliminate some interfering gases. This fluorescence is visualized by the
PM tube placed near the reaction chamber.
At the start of each "zero-ref", a shutter is placed between the UV lamp and the reaction chamber inlet
for 40 seconds. This electrical zero corresponds to the PM tube darkness current and the offset
voltage of the preamplifier, incorporated into the signal processing, it eliminates the possibility of drifts
with temperature and time.
The PM tube signal is amplified and is converted into digital values for processing by a microprocessor
that calculates the average of measurement values, checks the alarms and carries out monitor
operation diagnosis. These various values and information are displayed on an alphanumeric display
unit on the monitor front panel.
2–8
JULY 2002
AF22 MODULE
Duplication prohibited
2.4
Environnement S.A
SIMPLIFIED FLOW CHART OF PRINCIPAL PROGRAM
INPUTS/OUTPUTS INITIALIZATION MEMORY TEST
ALARM CHECK
ALARM DISPLAY
at end of warm-up
REFERENCE CYCLE
POSITION RANGE RELAY
PM SIGNAL ACQUISITION
Save this value if end od a "dark" cycle
UV SIGNAL ACQUISITION
Save this value if end of a "dark" cycle
ENERGY CORRECTION
MEAS-DARK - UV-ZERO
UV-ZERO
TOTAL VALUES (RT FUNCTION)
MEMORIZATION OF REFERENCE
REFERENCE
TEMPERATURE & ENERGY CORRECTION
MEASUREMENT-REFERENCE
CALCULATION
K SPAN
+ TEMPERATURE CORRECTION
MEAS. SO2 = SO2 X K SPAN
SCALING
If mg programmed
SO2(MG) = SO2(PPM) X CONV. FACT.
If offset programmed
ADD OFFSET
ACQUISITION OF MULTIPLEXER
PARAMETER AND PROCESSING
CONTROL OF PROGRAMMED PLC'S
RESULT
DISPLAY
JULY 2002
STORAGE
ANALOG
OUTPUTS
PRINTING
SERIAL LINK
2–9
AF22 MODULE
Environnement S.A
2.5
Duplication prohibited
AUTOMATIC RESPONSE TIME
In order to optimize its metrology, the AF22M monitor is equipped with a software function called
"automatic response time" which enables filtration of measurements depending on evolution of
concentrations of sulfur dioxide.
2.5.1
SIMPLIFIED PRINCIPLE OF OPERATION
An average of instantaneous readings is carried out corresponding to a minimum response time.
>MEAS@ AVERAGE
80
ª
º
«
¦1 >Pm signal @ »
» Reference zero
= « 16
«
»
« ¦ >UV detector signal @ »
¬ 1
¼
Then, a weighted average between the filtered values ( >MEAS@ FILTERED ) and the average
measurements ( >MEAS@ AVERAGE ) is recursively calculated according to the formula:
>MEAS@ DISPLAYED
>MEAS@ FILTERED (NEW)
X > MEAS@ FILTERED (OLD) Y [MEAS] AVERAGE
X +Y = 1
When the difference ( >MEAS@
FILTERED (OLD)
>MEAS@
AVERAGE
) exceeds a determined threshold, the
value of Y is increased, up to a maximum value of 0.98 which corresponds to a fixed response time of
TRMIN.
When >MEAS@
2.5.2
FILTERED (OLD)
>MEAS@
AVERAGE
is below the threshold, Y is progressively decreased.
PROGRAMMING THE RESPONSE TIME
The function of automatic response time may be activated or deactivated in the configuration Ö
measurement mode menu.
The minimum response time may also be modified in that menu.
See chapter 3 section 3.3.4.2 to obtain more information about programming of those functions.
2–10
JULY 2002
AF22 MODULE
Duplication prohibited
Environnement S.A
CHAPTER 3
OPERATING INSTRUCTIONS
3.1
3.2
INITIAL STARTUP
3–4
3.1.1
3.1.2
3–4
3–6
PROGRAMMING THE AF22M
3–7
3.2.1 SELECTION AND MODIFICATION OF THE PROGRAMMABLES PARAMETERS
3–7
3.2.2
3.3
PRELIMINARY OPERATIONS
STARTING UP THE UNIT
3.2.1.1
Screen areas definition
3–7
3.2.1.2
Definition of the most used functions of the keyboard
3–8
PROGRAMMING THE OPERATING PARAMETERS
3–8
3.2.2.1
Programming the digital parameters
3–8
3.2.2.2
Programming the configurable parameters with toggle list
3–8
DESCRIPTION OF THE DIFFERENT SCREENS
3–10
3.3.1
MAIN MENU
3–10
3.3.2
MEASUREMENT
3–11
3.3.3
3.3.4
3.3.2.1
MEASUREMENT Ö Instantaneous
3–11
3.3.2.2
MEASUREMENT Ö Average
3–14
3.3.2.3
MEASUREMENT Ö Synoptic
3–14
3.3.2.4
MEASUREMENT Ö Graphic
3–15
3.3.2.5
MEASUREMENT Ö Printout.
3–18
3.3.2.6
MEASUREMENT Ö Alarms display
3–18
SPAN
3–19
3.3.3.1
SPAN Ö Calibration
3–19
3.3.3.2
SPAN Ö Select Gas.
3–20
3.3.3.3
SPAN Ö Cycles
3–21
3.3.3.4
SPAN Ö Pressure
3-22
3.3.3.5
SPAN Ö E2Pot
3-22
CONFIGURATION
3–23
3.3.4.1
CONFIGURATION Ö Date/time/language
3–23
3.3.4.2
CONFIGURATION Ö Measurement mode
3–24
3.3.4.3
CONFIGURATION Ö Measure channels
3–25
3.3.4.4
CONFIGURATION Ö Offsets and units
3-25
3.3.4.5
CONFIGURATION Ö Alarm limits
3–26
3.3.4.6
CONFIGURATION Ö Analog outputs
3–26
3.3.4.7
CONFIGURATION Ö Analog inputs
3–27
3.3.4.8
CONFIGURATION Ö Relay and remote control
3-28
3.3.4.9
CONFIGURATION Ö Serial link
3-29
3.3.4.10 CONFIGURATION Ö Factory settings
JUNE 2004
3-29
3–1
AF22 MODULE
Environnement S.A
3.3.5
STORED DATA
3–30
3.3.6
TESTS
3–35
3.3.7
3.4
Duplication prohibited
3.3.6.1
TEST Ö Optical bench
3–36
3.3.6.2
TEST Ö MUX signals
3–37
3.3.6.3
TEST Ö Other commands
3–38
3.3.6.4
TEST Ö Serial link
3–38
3.3.6.5
TEST Ö ESTEL card
3–39
STOP MODE
3-40
CALIBRATION
3–41
3.4.1
OVERVIEW OF CALIBRATION AND CONCEPTS
3–41
3.4.1.1
Zero air generation
3-42
3.4.1.2
Span gas generation
3-42
3.4.1.3
Internal solenoid valves
3-43
3.4.2
3.4.3
3.4.4
3.4.5
CHECK OF ZERO AND SPAN POINT
3–43
3.4.2.1
Facilities required
3-43
3.4.2.2
Procedure
3-43
3.4.2.3. Use of automatic cycles
3-44
SPAN ADJUSTMENT
3–44
3.4.3.1
Facilities required
3-44
3.4.3.2
Procedure
3-44
3.4.3.3
Use the AUTO-CAL automatic cycle
3-45
MULTI-POINT CALIBRATION
3–46
3.4.4.1
Overview
3-46
3.4.4.2
Equipment required
3-46
3.4.4.3
Procedure
3-46
INTERNAL PERMEATION OVEN (OPTIONAL)
3–48
Figure 3-1 - Electrical connections
Figure 3-2 - Fluids connection
Figure 3-3 – Permeation tube installation
Figure 3-4 – Software overview
Figure 3-5 - Printout example
Figure 3-6 - Pressurized gas connection example
Figure 3-7 - Typical calibrator
3–3
3–4
3–5
3–9
3–34
3–42
3–45
Table 3–1 - DB37 and DB25 connectors links
Table 3–2 - MUX signals (Acceptable limits on the multiplexer 1 to 16 channels)
3–3
3–37
3–2
MAY 2007
AF22 MODULE
Duplication prohibited
3
Environnement S.A
OPERATING INSTRUCTIONS
Figure 3-1 - Electrical connections
Table 3-1 - DB37 and DB25 connectors links
RS232 / 422 serial links
COM1
2 - TX
3 - RX
4 - RTS
7 - GND
20 - DTR
21 - TX
11 - RX
COM2
14 - TX
16 - RX
7 - GND
ESTEL BOARD(S)
PIN N°
1
2
3
4
5
6
7
8
9-28
10-29
11-30
12-31
13-32
14-33
15
16
NOTE :
CONNECTION
+ ANA OUTPUT 1
+ ANA OUTPUT 2
+ ANA OUTPUT 3
+ ANA OUTPUT 4
+ANA INPUT 1
+ANA INPUT 2
+ANA INPUT 3
+ANA INPUT 4
RELAY 6 CONTACT
RELAY 5 CONTACT
RELAY 4 CONTACT
RELAY 3 CONTACT
RELAY 2 CONTACT
RELAY 1 CONTACT
REMOTE CONTROL 1
REMOTE CONTROL 2
PIN N°
17
18
19
20
21
22
23
24
25
26
27
34
35
36
37
CONNECTION
REMOTE CONTROL 3
REMOTE CONTROL 4
+5VCC
ANA OUTPUT GROUND
ANA OUTPUT GROUND
ANA OUTPUT GROUND
ANA OUTPUT GROUND
ANA INPUT GROUND
ANA INPUT GROUND
ANA INPUT GROUND
ANA INPUT GROUND
REMOTE CONTROL GROUND
REMOTE CONTROL GROUND
REMOTE CONTROL GROUND
REMOTE CONTROL GROUND
Output relays contacts are normally open and potential free.
Remote controls are made by closing a potential free dry contact
Analog inputs accept maximum 2.5 VCC.
JUNE 2004
3–3
AF22 MODULE
Environnement S.A
3.1
Duplication prohibited
INITIAL STARTUP
The monitor is checked and calibrated in the factory before delivery.
3.1.1
PRELIMINARY OPERATIONS
Start-up first consists in carrying out the following preliminary operations:
– Visually examine the interior of the instrument in order to ensure that no element has been
damaged during transport.
– Remove the caps from the "gas" inlets and outlets on the unit (keep these aside for future
storage, see Chapter 1.2.3).
WARNING : Do not remove zero inlet cap if built-in zero filter is present
Connect the 4/6 Teflon air sampling tube to the "sample inlet", after having checked for the
presence of a Teflon filtering membrane in the inlet dust filter (fig.3-2)
Connect the digital outputs to the DB25 connector (see Table. 3-1).
Connect the analog inputs / outputs to the DB37 connector(s) (see Table 3-1).
– Connect the mains power supply cable to a socket: 230 V, 50 Hz + ground or 115 V, 60 Hz
+ ground depending on the voltage specified on order.
Zero air inlet of the
permeation bench
(option)
Figure 3-2 - Fluids connection
3–4
MAY 2007
AF22 MODULE
Duplication prohibited
3.1.2
Environnement S.A
PERMEATION TUBE INSTALLATION (WHEN OPTION AVAILABLE)
2
3
1
Figure 3-3 – Permeation tube installation
Pull the Teflon tap (1) of permeation oven inlet (2). After putting it out of its housing, insert permeation
tube (3) in permeation oven, with porous diaphragm faced to the bottom.
NOTE : It is essential to NOT OPEN permeation tube and to NOT PERFORATE the porous
diaphragm.
If analyzer has to be kept switched off for a while, it is necessary to remove tube from permeation
oven, place it in its delivery box with drying bags, and store it in a cool place.
JUNE 2004
3–5
AF22 MODULE
Environnement S.A
3.1.3
Duplication prohibited
STARTING UP THE UNIT
Press the ON/OFF switch located on the front panel. The monitor goes into the "warm-up" cycle. The
duration of this cycle is a function of the time elapsed since the last switch off.
The warm-up cycle is terminated when the two following conditions are satisfied:
– All metrological parameters are within operational limits,
– The analyzer has made 10 consistent measurements within ± 4 ppb.
Display at start-up: the WARM UP message appears in the top right corner.
Display after warm-up : the measurement display mode can be chosen in CONFIGURATION Ö
Measurement mode. Example given here below: Synoptic screen.
STOP
After some time (programmable in CONFIGURATION Ö Measurement mode)
without action on any key, the screen passes in stand-by mode.
Pressing down any key makes it going back to display mode.
3–6
MAY 2007
AF22 MODULE
Duplication prohibited
Environnement S.A
3.2
PROGRAMMING THE AF22M
3.2.1
SELECTION AND MODIFICATION OF THE PROGRAMMABLE PARAMETERS
The keyboard is located under the LCD. The bottom line gives the function of each key for the current
screen.
The title of the menus and the selected fields are displayed in reverse video. By default the first line of
the menus is selected. In the next paragraphs, the selected fields are symbolized in white on black
background.
3.2.1.1
Screen areas definition
1
2
3
Information area: displays the date and time in the top left corner. In the top right corner, the
WARM UP, REFERENCE or SPAN messages blinks. The ALARM message appears if an
operating fault is detected within the instrument operating parameters.
1
2
Measurement or configuration area: displays the measurement parameters (gas, value,
units ...) or the programmable parameters according to the selected menu.
3
Status areas and keys functions: displays the keys functions, the analyzer operating mode
and the SO2 inlet ("sample" in the example above).
NOTE : In the next paragraphs, the keys are symbolized by the icon or function displayed inside a
rectangle.
JUNE 2004
3–7
AF22 MODULE
Environnement S.A
3.2.1.2
Duplication prohibited
Definition of the most used functions of the keyboard
(The availability of these functions is context dependent)
Used to display the previous menu or to abort the current operation (parameter programming,
etc.)
Used to select the required sub-menu and the parameter to be modified. It is also used to
increase the digit whose modification is in progress.
Used to select the required sub-menu and the parameter to be modified. It is also used to
decrease the digit whose modification is in progress.
Moves the cursor to the left (only available during numerical parameters modifications).
Moves the cursor to the right (only available during numerical parameters modifications).
Authorizes the selected parameter modification.
Used to valid the selection or the parameter value whose modification is in progress.
Print
It is used to print out the current screen.
>> Used to display the next page. When there are several parameters, pressing down this key
allows to display the next parameters.
3.2.2
PROGRAMMING THE OPERATING PARAMETERS
3.2.2.1
Programming the digital parameters
Select the parameter with the
key in the appropriate menu, press down the
or
key to
st
access to the modification of the parameter, the 1 digit blinks. Select the digit to be modified with the
or
key then increase it with the
key or decrease it with the
validates the modifications of the selected field, the
key. The
key
key cancels the modifications of the selected
field.
3.2.2.2
Programming the configurable parameters with toggle list
Select the parameter with the
or
key in the appropriate menu, press down the
access to the modification of the parameter, the field blinks. Select with the
wanted value in the toggle list. The
or
key to
key the
key validates the modifications of the selected field, the
key cancels the modifications of the selected field.
3–8
JUNE 2004
AF22 MODULE
Duplication prohibited
Environnement S.A
MAIN MENU
Measurement
Span
Configuration
Stored Data
Tests
Instantaneous
Calibration
Date/Time/
Language
Selection
Optical bench
Select gas
Measurement
Mode
Tablular
MUX signals
Cycles
Measure
channels
Histogram
Other
commands
Pressure
Offsets and
units
Printer output
Serial link
Stop Mode
Cycle
Average
Cycle
Synoptic
Cycle
Graphic
Alarm limits
Menu
Speed
Base
Scale
Printout
Alarm display
Estel card *
Analog output *
Analog input *
Relays and
remote control
Serial link
Factory settings
Figure 3-4 – Software overview
*: appears in the menu when option present
JUNE 2004
3–9
AF22 MODULE
Environnement S.A
3.3
DESCRIPTION OF THE DIFFERENT SCREENS
3.3.1
MAIN MENU
Duplication prohibited
This screen is used to choose the menus giving access to the analyzer operating parameters
Select the menu with the
Example:
ACTION
or
key, validates the selection with the
DISPLAY
key.
REMARKS
– Displays the main menu, the 1ST item
is selected by default.
– Selects the next item.
– Selects the next item.
– Validates the selection (Configuration
menu) and displays the sub-menu.
– Goes back to the previous menu.
NOTE : To make the reading easier, when a sub menu is quoted in the text, the corresponding menu
is reminded before (ex. CONFIGURATION Ö Date / time / language).
3–10
JUNE 2004
AF22 MODULE
Duplication prohibited
3.3.2
Environnement S.A
MEASUREMENT
This screen is used to choose the measurement display mode: instantaneous, average, synoptic
(diagram) or graphic, to activate the real-time printout and to display possible alarms.
3.3.2.1
MEASUREMENT Ö Instantaneous
Definition of the specific keys to this screen
Sample
Zero
Selects the sample gas inlet. The gas is continuously sampled through the inlet dust filter. The
measurement mode, the unit and the range are those chosen in the Configuration menu and in
the corresponding sub-menus. This mode can be interrupted at any time by starting an automatic
cycle or by manual selection of another mode or gas inlet (zero or span).
Selects the zero gas inlet. Allows manual control of the zero on the built in zero filter or on the
optional external zero gas inlet. The analyzer gives its reading on zero gas (possibly increased
by a programmed offset).
This operation allows to check the stability and the drift of the zero reading of the analyzer to
determine the necessity to launch a reference cycle or to program its repetition period.
JUNE 2004
3–11
AF22 MODULE
Environnement S.A
Span
Duplication prohibited
Selects the span gas inlet. Allows manual control of the span. The analyzer gives its reading on span
gas (possibly increased by a programmed offset).
This operation allows to check the stability and the drift of the span reading of the analyzer to determine
the necessity to launch an auto-span cycle or to program its repetition period.
>>
Cycle
Used to display the next page. When there are several parameters, pressing down this key allows to
display the next parameters.
Gives access to the screen allowing manual launching of cycles.
Definition of the specific keys to this screen
Z.ref
Escape
End
Allows to manually launch an automatic corrective cycle in case of electrical zero and zero
gas difference. At the top right corner, zero reference and the gas inlet on which the cycle
is conducted are displayed.
Allows to skip the current cycle without computing the new Zero-reference.
Allows to manually shorten the cycle duration.
Be careful : the new zero-reference displayed value is computed, this could occur a wrong
zero reference value.
3–12
JUNE 2004
AF22 MODULE
Duplication prohibited
Auto
Environnement S.A
Allows to launch manually an automatic span cycle. The analyzer adjusts automatically its
span factor K in order to equal its reading value (minus the programmed offset) and span
gas concentrations. The span gas concentrations are programmable in the
SPAN Ö Select gas menu, the concentration programmed for the gas inlet used for
calibration is reminded in the top right corner of the screen (CAL=XXXX). The adjustment
takes place for the duration programmed in the SPAN Ö Cycles menu. The cycle time
duration countdown is displayed in the right top corner of the screen, the cycle is achieved
when count is 0000 sec. Cycle can be shorten pressing down the
Auto
key. The new span
factor is then memorized if reading value equals + 5 % span concentration.
Escape
End
Allows to skip the current cycle without computing the new K-Span.
Allows to manually shorten the cycle duration.
Be careful : the span factor is computed even if the read and the target values are not
equal. This could occur a wrong span factor.
This function launches the auto-span using the gas inlet selected before pressing
STOP
down the
Auto
key, select the correct gas inlet before executing an auto span
To go back to normal measurement after manual auto-span, press down the
Sample
key to select the
sample inlet again.
JUNE 2004
3–13
AF22 MODULE
Environnement S.A
3.3.2.2
Duplication prohibited
MEASUREMENT Ö Average
Definition of the specific keys to this screen
Sample
3.3.2.3
Zero
Span
Cycle
have the same function as the screen MEASUREMENT Ö Instantaneous.
MEASUREMENT Ö Synoptic
This screen represents the flow circuit and displays significant operation parameter values: Gas,
concentration and unit (1), PM high voltage (2), PM signal (3), optical bench temperature (4), Gas
pressure (5), internal temperature (6), UV detector signal (7), voltage proportional to UV lamp current
(8).
6
5
7
4
8
2
3
1
Definition of the specific keys of this screen
Sample
Zero
Span
3–14
Cycle
Have the same function than the screen MEASUREMENT Ö Instantaneous.
MAY 2007
AF22 MODULE
Duplication prohibited
3.3.2.4
Environnement S.A
MEASUREMENT Ö Graphic
This screen is used for graphic plotting of the measurement values on sample or zero/span gas inlet.
The vertical line shows the current position: The refreshed measurements are given on the left side of
this line. The vertical full scale of the graphic is the one programmed for the analog outputs.
Definition of the specific keys to this screen
Sample
Zero
Span
have the same function as the screen MEASUREMENT Ö Instantaneous.
When sample inlet is already active, pressing down the
Sample
key refreshes the graph. Press down the
>> key to select the parameter to display.
3.3.2.4.1
Graphic Ö "Menu" screen
Pressing down the
Menu
x
Plotting speed
x
Base line
x
Full scale
The
R.S.T
key gives access to the following graphic adjustments:
key is used to reset the graph to zero.
Plotting speed
Current scale
Graphic base
JUNE 2004
3–15
AF22 MODULE
Environnement S.A
3.3.2.4.2
Duplication prohibited
Graphic Ö screen "Base"
Pressing down the
Base
key allows to adjust the value of the graph base line (the minimum value is
zero, the maximum value is just inferior to the full scale).
__
_
Selects the inferior base line among 5000, 2000, 1000, 500, 200, 100, 50, 20, 10, 5, 2, 1, 0.
+
Selects the superior base line among 0, 1, 2, 5, 10, 20, 100, 200, 500, 1000, 2000, 5000.
++
3.3.2.4.3
Divides 10 times the current base line (when the base line is 5, it resets to zero)
Multiplies 10 times the current base line.
Graphic Ö screen "Speed"
Pressing down the
Speed.
key allows to adjust the plotting speed on the screen (the minimum value is 1
second, the maximum value is 60 seconds).
3–16
JUNE 2004
AF22 MODULE
Duplication prohibited
- 10s
Decreases 10 seconds to the current plotting speed
- 1s
Decreases 1 second to the current plotting speed.
+1s
Increases 1 second to the current plotting speed.
+10s
Environnement S.A
Increases 10 seconds to the current plotting speed.
The programmed time is the interval duration between 2 points of the graph
E.g.: when the plotting speed is 10 seconds, the graphic screen lasts for 240 x 10 = 2400 seconds.
3.3.2.4.4
Graphic Ö screen "Scale"
Pressing down the
Scale
key allows to adjust the full scale of the graph (the minimum value is just
superior to the base line, the maximum value 10000)
This screen allows to adjust the full scale of the graphic (minimum not exceeding the basis line,
maximum value reaching 1000).
__
Selects the current scale among 5000, 2000, 1000, 500, 200, 100, 50, 20, 10, 5, 2, 1, 0.
+
Selects the current scale among 0, 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000, 5000.
++
JUNE 2004
Divides 10 times the current scale (when the scale is 5, it reset to zero.)
_
Multiplies 10 times the current scale.
3–17
AF22 MODULE
Environnement S.A
3.3.2.5
Duplication prohibited
MEASUREMENT Ö Printout.
This menu is used to start real time printout on a serial printer connected to the COM2 serial port. It is
also used to define measurement average calculation period and printing rate from (0001 to 9999mn).
3.3.2.6
MEASUREMENT Ö Alarms display
This screen displays the operating faults in case of alarm. Possible corrective actions for these faults
are given in chapter 5.
3–18
JUNE 2004
AF22 MODULE
Duplication prohibited
3.3.3
Environnement S.A
SPAN
This menu gives access to the followings functions:
x
Programming the calibration factors K.
x
Programming the span gas values.
x
Gas inlet selection for span cycles.
x
Programming the period and duration of the automatic cycles.
x
Pressure sensors calibration
Programmable digital potentiometer (E2Pot) allows to:
3.3.3.1
x
Adjust mercury lamp current and consequently measurements and reference signal amplitude;
x
Adjust flow rate control point.
SPAN Ö Coefficients
This screen is used to modify manually calibration factors. The factors variations after a new auto-cal
cycle are displayed in the "Delta %" fields. To reset manually the Delta % in case of calibration alarm
due to a wrong use of the auto-cal function, select the Factor field of the gas whose the Delta % is
greater than 5.0, press down the
key and press down the
key and the
key. Exit the screen by pressing down the
key to select again the SPAN Ö Calibration screen in order to
refresh the Delta % display.
JUNE 2004
3–19
AF22 MODULE
Environnement S.A
3.3.3.2
Duplication prohibited
SPAN Ö Select Gas.
This screen is used to associate a span gas concentration to each gas inlet. These concentrations are
the reference values for manual or automatic auto-calibration cycles.
3–20
JUNE 2004
Duplication prohibited
3.3.3.3
AF22 MODULE
Environnement S.A
SPAN Ö Cycles
This screen is used to program the period and the duration of the automatic cycles, the durations
programmed here are also those of the manually launched cycles.
The possible automatic cycles are:
ZERO : zero air check
SPAN : span gas check.
Z.Ref : automatic zero adjust correction
AUTO : automatic span factor correction
The “Remote” fields are used to configure the cycles remote controls (ESTEL board option), ZERO,
ZERO REFERENCE, AUTO or SPAN. The status programmed in the “Cyclical” fields (ON = active,
OFF = inactive) governs the analyzer reaction when a dry contact is closed on the remote controls
inputs (see Table 3.1).
The “Inlet” fields allow to select gas inlets used during the automatic sequences. The reference
concentrations for automatic calibration are those programmed in the previous menu.
The “Starting time” field is used to program the hour when the cycles are launched. If a 24 h ZERO
cycle, a 24 h AUTO cycle, and a 24 h SPAN cycle are programmed, the sequence is launched at
starting time according to the following priority: Z.Ref., ZERO, AUTO, then SPAN.
To inhibit an automatic cycle, program 0000h in the “Period” field. To inhibit both an automatic and a
manual cycle, program 0000s in the “Timing” field.
JUNE 2004
3–21
AF22 MODULE
Environnement S.A
3.3.3.4
Duplication prohibited
SPAN Ö Pressure
This screen is used to program the calibration curve of the pressure sensors.
Pressure sensors calibration:
Connect a reference pressure sensor in parallel to the pressure sensor to be calibrated.
Enter the slope value (A) and its intercept (B) in the pressure calibration window fields.
3.3.3.5
SPAN Ö E2Pot
This screen allows to adjust, in a digital way, electric gains on measurement signals.
+
Pressing down this key increase the signal
-
Pressing down this key decrease the signal
3–22
MAY 2007
Duplication prohibited
3.3.4
3.3.4.1
AF22 MODULE
Environnement S.A
CONFIGURATION
This menu gives access to the following functions:
x
Response time programming.
x
Dilution function programming.
x
Analog outputs configuration.
x
Unit change and offset adjustment
x
Alarm thresholds, activation and assignment of the alarm relays.
x
Serial link programming.
x
Reset of the main programmable parameters.
CONFIGURATION Ö Date/time/language
This screen is used to set the internal clock of the analyzer, as well as to choose the displayed
language among French, English, German, Italian and Spanish. It also shows the software version
number to remind in case of software dysfunction.
MAY 2007
3–23
AF22 MODULE
Environnement S.A
3.3.4.2
Duplication prohibited
CONFIGURATION Ö Measurement mode
This screen is used to program the response time from about 20 sec. to 120 sec. The first digit of the
programmed number activates (1) or de-activates (0) the automatic response time function, the
second item divides the basic electronic integration time (120 sec.)
Examples: Response time = 13 gives an automatic response time with a minimum of 40 sec.
Response time = 03 gives an integration time fixed to 40 sec.
The advised value is; Response time =11
(see chapter 2, automatic response time)
This screen gives also access to the dilution function: to measure very high concentrations (i.e.
founded in industrial environment), it is necessary to bring them to values corresponding to the range
of analyzer by inserting a dilution system in the sampling line to get:
C analyser inlet
C sample
Kdilution
The real concentration display is obtained by applying a K Dilution multiplying factor to the measured
concentration.
"Latch DAC" field: when this field is ON, it latches the analog outputs to the last measurement values
during the zero or span cycles in order not to perturb eventual data loggers.
"Maintenance" field: when this field is on, it allows to trigger one of the alarm relays (see section
3.3.4.5 and table 3.1). The maintenance mode condition is reminded on the MEASUREMENT
screens.
"Starting screen" field allows to choose the screen displayed after warm-up when starting on the
analyzer. 4 choices are offered: Instantaneous, Synoptic, Average, Graphic, corresponding to the
menu MEASUREMENT.
Screen std –by delay" field allows to program the time delay after which, without any action on
keyboard, the screen passes in stand-by mode.
"Zero filter life time" field allows to program a day down-counter which triggers, when zero, a filter
alarm message. The setting value depends on the analyzer using conditions. The factory setting
value, 365 days, corresponds to the advised maintenance frequency.
3–24
JUNE 2004
Duplication prohibited
3.3.4.3
AF22 MODULE
Environnement S.A
CONFIGURATION Ö Measure channels
This screen is used to select the parameter, the display format and the unit for each measure channel.
The measure channels programming allows to display (MEASUREMENT Ö Instantaneous screen or
MEASUREMENT Ö Average screen) and to store (STORED DATA menu) other parameters than the
one displayed by default (SO2). It allows to store MUX channels and analog inputs (ESTEL option).
x
The 8 "Channels" fields are used to choose the parameter among: SO2, H2S (if H2S rack option
available), TRS (if TRS rack option available), GND, Int.t°, Opt.t°, Aux.t°, Flow r., Pressure,
+15V, -15V, I.Pbse, I.IR, Signal.
x
The "Formats" fields are used to choose the display format among 4 possibilities (X.XXX,
XX.XX, XXXX.X, XXXX). "Auto" manages the comma in order to display the best resolution at
any time.
The "Units" fields refer to the units programmed in screen CONFIGURATION Ö Offsets and units, or
CONFIGURATION Ö Analog inputs.
3.3.4.4
CONFIGURATION ÖOffsets and units
This screen is used to program the offset. This value is added to the measurements. It is also used to
program the conversion factors from ppm to mg/m3, when the mg/m3 unit is selected. The conversion factor
of 2,660 is applied for normal conditions of temperature and pressure, as follow : (20°C and 101,3 kPa).
MAY 2007
3–25
AF22 MODULE
Environnement S.A
3.3.4.5
Duplication prohibited
CONFIGURATION Ö Alarm limits
2 limits are programmable for the programmed parameter: Threshold 1 and Threshold 2, allowing to
activate relays and alarm messages. When the "Alarms display" field is “OFF”, displays and alarm
relays are inhibited.
3.3.4.6
CONFIGURATION Ö Analog outputs
This screen allows to choose the analog outputs parameters (only when ESTEL board option is
available) among:
x
The SO2 concentration, the H2S concentration (optional), the TRS concentration (optional).
x
From MX01 to MX16, the 16 channels multiplexer
x
The external inputs.
The chosen parameters correspond to the analog outputs.
This screen is used to program the ranges for each displayed parameter. The ranges correspond to
the analog output full scale.
Scale 1 corresponds to the analyzer standard range. The analyzer switches to scale 2 when scale 1 is
exceeded. It switches again to scale 1 when measurement decreases below 85 % of the full scale 1.
This screen is also used to choose the parameters units among ppb, mg/m3, mV, °C or hPa.
3–26
JUNE 2004
Duplication prohibited
3.3.4.7
AF22 MODULE
Environnement S.A
CONFIGURATION Ö Analog inputs
This screen is used to program the analog inputs characteristics.
JUNE 2004
x
"ESTEL card" field allows to select the board to be programmed: each ESTEL board has 4
analog inputs.
x
"Name" fields allow to enter 8 alphanumeric digits.
x
"Unit" fields allow to choose a unit among : none, ppt, ppb, ppm, μg/m3, mg/m3, gr/m3, μg/Nm3,
mg/Nm3, gr/Nm3, μg/Sm3, mg/Sm3, gr/Sm3, %, μgr, mgr, gr, mV, U, °C, °K, hPa, mb, b,l, Nl, Sl,
m3, l/min, NI/min, Sl/min, m3/h, Nm3/h, Sm3/h, m/s or km/h, in a scrolling menu.
x
"aX + b" fields allow to enter the calibration curve for each parameter.
x
"Meteo" fields allow to assign the channel where meteorological parameters are connected in
order to apply a trigonometric treatment to those data.
3–27
AF22 MODULE
Environnement S.A
3.3.4.8
Duplication prohibited
CONFIGURATION Ö Relays and remote control
This screen allows to configure the function of each input / output of the Estel board(s).
x
"Estel card Nb" field allows to choose what board to configure.
x
"Relays" fields allow to control the relays according to the following situations :
Disable
Ö Relay not assigned
General alarm
Ö Any operating fault triggers the relay
Ch.1 > Thrs.1
Ö Limit 1 channel 1 exceedance triggers the relay
Ch.1 > Thrs.2
Ö Limit 2 channel 1 exceedance triggers the relay
Ch.2 > Thrs.1
Ö Limit 1 channel 2 exceedance triggers the relay
Ch.2 > Thrs.2
Ö Limit 2 channel 2 exceedance triggers the relay
Ch.3 > Thrs.1
Ö Limit 1 channel 3 exceedance triggers the relay
Ch.3 > Thrs.2
Ö Limit 3 channel 3 exceedance triggers the relay
Over range
Ö Range 2 exceedance triggers the relay
Flow rate
Ö Abnormal flow rate triggers the relay
Temperature
Ö Abnormal temperature in the analyzer triggers the relay
Pressure
Ö Barometric pressure in chamber
Null gas
Ö On zero, relay is triggered
Span
Ö On span, relay is triggered
Ref-Zero
Ö On Ref-Zero, relay is triggered
Auto Span
Ö On Auto Span, relay is triggered
Warm-up
Ö On Warm-up, relay is triggered
Stand-by
Ö On Stand-by, relay is triggered
Maintenance
Ö Relay triggered when the analyzer is in maintenance mode
x
"Type" fields allow to control (NC) or not (NO) the relays when alarm OFF.
x
"Mode" field allows to configure working remote controls mode.
Two different modes are possible:
"State" mode: control is activated as long as remote control is active. When remote control falls down,
control is no more active.
"Rise" mode: control is activated when remote control activation is detected. When it is down, control
remains active. It is necessary to re-activate the same remote control to make down the previous
remote control.
3–28
JUNE 2004
Duplication prohibited
3.3.4.9
AF22 MODULE
Environnement S.A
CONFIGURATION Ö Serial link
This screen is used to configure the Serial links (COM 1 and 2).
The baud rate, format and communication mode of the 2 channels are programmable among:
3.3.4.10
x
Baud rate: 1200, 2400, 4800, 9600, 19200, 38400 (limited to 19200 bds at present time)
x
Format : 7n1, 7o1, 7e1, 7n2, 7o2, 7e2, 8n1, 8o1, 8e1, 8n2, 8o2, 8e2
x
Communication mode: Mode 4, impress. to send measurements to printer in real time, Jbus,
Special1, and Special2.
CONFIGURATION Ö Factory settings
When this item is selected, pressing down the
JUNE 2004
key displays the screen shown here below:
3–29
AF22 MODULE
Environnement S.A
3.3.5
Duplication prohibited
STORED DATA
The access to stored data management is directly done from Main Menu. The stored data consists in
the average of analyzer measurements within a defined time interval.
1
This screen allows to parameter data recording period from 1 to 1440 min (i.e. 24 hours) and informs
about memory status:
Free memory: from 80 Ko in standard operation, it can be increased to 464 Ko in adding a 384 Ko
memory board (optional). This board is automatically detected when switching on the analyzer and it is
indicated on the screen (1).
Storage : it is the possible records number, it depends on free memory
Autonomy: it is the duration (days, months, years, hours, minutes) while memory can store data,
considering free space and data recording period. In the here-above example: 4 days,
0 month, 17 hours, 46 minutes.
Data can be edited in the form as table or histogram: this screen allows to program date and hour of
edition beginning, date and hour of edition end, histogram column width.
Menu
key gives access to table or histogram data display, printing functions, and memory reset-to-zero.
3–30
JUNE 2004
AF22 MODULE
Duplication prohibited
Environnement S.A
Stored data edition in the form as tabular
This screen presents stored data list according to parameters defined in the screen before. The
running mode (measurement, zero, calibration…), during a memorization period, is coded in the status
column. The status codes meaning are:
00
Measurement valid
01
Range 2 over shooting
02
General alarm
04
Calibration fault
08
Zero measurement
10
Span measurement
20
Maintenance
40
Less than 2/3 of valid measurements during the average period
80
Power supply failure
FF
Configuration modification
The displayed status code corresponds to the summation of the status codes (hexadecimal numbers)
that occurs during the memorization period.
Example: with an average period of 20 min:
5 min zero and 15 min measurement give the 00 status code and the displayed average is the 15 min.
measurement average.
11 min. zero and 9 min. measurement give the 08 status code and the displayed average is the 11
min. zero average.
Definition of the specific keys to this screen:
Select the previous or the next page.
<
Select stored data beginning or end.
Display the other measure channels if more than 3 channels are programmed in
CONFIGURATION Ö Measure channels screen
JUNE 2004
3–31
AF22 MODULE
Environnement S.A
Duplication prohibited
Stored data edition in the form as histogram
This screen displays records in the form as columns; each column corresponds to the measurements
average within the data-recording period as defined in STORED DATA screen. Only one channel is
displayed at once. The information line gives first record date and hour, the channel name, and,
alternatively blinking, full scale with unit, and data recording period.
This area
corresponds
to no stored
data.
Definition of the specific keys to this screen
Return to previous menu.
Display previous stored data plotting.
Display next stored data plotting.
X 2 zoom in
1/2 zoom out.
>> Select the next measure channel, when more than one measure channel is programmed.
3–32
JUNE 2004
Duplication prohibited
AF22 MODULE
Environnement S.A
Stored data printing
To print data, press down the
Print
key found in "Menu" function of "Memorized data" screen. The
blinking message "Printing…" indicates printed data output. Data printing can be suspend at any time,
pressing down F1 key. When printing is finished, the screen displays the message "Printing finished".
When none communication port is programmed on printer output (serial port), the error message
"Printing not set" is displayed.
JUNE 2004
3–33
AF22 MODULE
Environnement S.A
Duplication prohibited
AF22M[3.1]
11-10-2001
SO2
HH:MM status PPB
10:15 00
14.7
10:30 00
21.7
10:45 00
21.4
11:00 00
17.9
11:15 00
16.0
11:30 00
14.7
11:45 00
13.9
12:00 00
13.5
12:15 08
1.0
12:30 00
12.9
12:45 00
12.7
13:00 00
12.3
13:15 00
11.8
13:30 00
10.6
13:45 00
9.2
14:00 00
8.5
14:15 00
8.0
14:30 00
7.7
14:45 00
6.8
15:00 00
7.6
15:15 00
7.5
EXT1
hPa
1001.7
1001.4
1002.1
1002.6
1002.4
1000.7
1001.4
1002.1
1001.8
1001.6
1001.1
1000.8
1001.0
1001.3
1001.5
1001.8
1002.1
1002.3
1002.5
1002.6
1002.4
MX13
ºC
39.4
39.5
39.4
39.4
39.4
39.4
39.4
39.5
39.4
39.3
39.2
39.1
39.0
38.9
38.7
38.7
38.5
38.3
38.2
38.1
38.0
15:45
16:00
16:15
16:30
16:45
17:00
17:15
17:30
17:45
1002.4
1002.3
1002.6
1002.6
1002.5
1001.8
1002.4
1002.2
1002 0
37.9
37.5
37.5
37.7
37.8
37.9
38.0
37.9
37 9
00
00
00
00
00
00
00
00
00
8.1
8.5
8.5
8.7
9.1
9.2
9.2
9.0
9 1
Figure 3-5 - Printout example
Memory reset to zero
Pressing down the
Reset
key allows to empty storage memory. This action is irreversible: before to
do it, the software asks you to confirm. If your answer is "YES", the software resets end edition dates
and hours to the current dates and hours.
3–34
JUNE 2004
AF22 MODULE
Duplication prohibited
3.3.6
Environnement S.A
TESTS
This screen gives access to the following functions:
JUNE 2004
x
Optical and flow parameters checking when maintenance operations occurs.
x
Serial link checking.
x
Checking of the ESTEL board working (when option available)
3–35
AF22 MODULE
Environnement S.A
3.3.6.1
Duplication prohibited
TESTS Ö Optical bench
This screen is used to follow-up the measurement parameters periodically or occasionally.
x
Pm = instantaneous amplified PM signal
x
Pm0 = PM black signal (shutter closed) memorized during the last reference cycle
x
UV = Instantaneous amplified UV detector signal
x
uv0 = UV detector black signal (shutter closed) memorized during the last reference cycle
x
brut = UV energy compensated raw measurement
x
REFER = "brut" memorized during the last reference cycle
x
inst = instantaneous measurement in ppb
x
moy = last memorized measurement average
x
Optical T° = optical bench temperature
x
Internal T° = internal temperature of the analyzer
x
B. Perm. T° = Permeation bench temperature (optional)
x
Flow rate = Voltage proportional to flow rate
x
Pressure = vacuum pressure in the chamber.
x
Lamp I = lamp current
Definition of the specific keys of this screen
Sample
Zero
Span
have the same function as the screen Measurement Ö Instantaneous.
3–36
MAY 2007
AF22 MODULE
Duplication prohibited
3.3.6.2
Environnement S.A
TESTS Ö MUX signals
This screen is used to check the multiplexer signals.
NOTE : The "XXXX mV" displayed value will be checked according to the acceptable limits in next table.
Table 3–2 - MUX signals (Acceptable limits on the multiplexer 1 to 16 channels)
Channel
1
Display
GND
Parameters
Analog ground
Lower Limit
0 mV
2
Int. T°
Internal temperature of the analyzer
3
Opt. T°
Temperature of optical chamber
4
PM HV
5
Flow r.
Voltage proportional to high
voltage applied to photo
multiplier tube (***)
Flow rate sensor voltage.
6
Pressure
7
100 mV
Normal
0 mV
Upper limit
50 mV
300 mV
600 mV
450 mV
500 mV
2400 mV
3000 mV
1000 mV
2800 mV
3500 mV
3300 mV
3200 mV
4000 mV
5000 mV
Pressure of measurement chamber (*)
1200 mV
4000 mV
5000 mV
-15V
Test point of power supply voltage
1200 mV
1500 mV
1600 mV
8
+15V
Test point of power supply voltage
1200 mV
1500 mV
1600 mV
9
UV Sig
7000 mV
9000 mV
10
-------
Measurement of UV lamp signal used to
1000 mV
compensate for any possible drift.
Not used
–
11
PM sig.
Measurement signal at PM amplifier
12
UV cur.
13
H2S T°
TRS T°
14
B.P. T°
15
O2
Voltage proportional to temperature of
2200 mV
built-in permeation oven (**)
_
O2 sensor (**)
16
2.5 V ref.
Reference voltage of analog / digital converter
–
Embase
HAMA
SDS
–
–
0 mV
–
9900 mV
Lamp current
250 mV
380 mV
440 mV
Thermocouple probe temperature (***)
14 mV
0 mV
17 mV
100 mV
22 mV
500 mV
2234 mV
2260 mV
2480 mV
_
_
2500 mV
2520 mV
(*) These values are given as information and they depends on the A and B linearization coefficients
(SPAN Ö Pressure screen)
(**) Only when the corresponding options are available in the analyzer.
(***) According to the PM embase (HAMA. or SDS) specified in the Check List attached to the analyzer.
JANUARY 2008
3–37
AF22 MODULE
Environnement S.A
3.3.6.3
Duplication prohibited
TESTS Ö Other commands
STOP
The selection of this menu makes ineffective some commands and regulations.
When the instrument is again in measurement mode, some alarms could occur.
This screen is used to check that the MODULE board is working correctly.
Jxx refers to the connector command, SV refers to solenoid valve.
3.3.6.4
TESTS Ö Serial link
This screen allows to check the serial link and shows the inputs / outputs which have to be strapped
when the serial link is not connected and you still want to check it are:
2-3: Emission/Reception, 4-6 and 7-8: modem signals.
3–38
JUNE 2004
AF22 MODULE
Duplication prohibited
3.3.6.5
Environnement S.A
TESTS Ö ESTEL card
This screen is only displayed when the option is available.
It is used to set the analog outputs and to monitor the working state of the remote controls and analog
inputs.
The “Estel card Nb:” field is used to select the board to be tested.
The “DA.C” (digital to analog converter) fields are used to program the number of points generated at
analog output.
The “Ax + B” fields are used to program the span factors of each output. These factors are calculated
according to the value measured at the output.
The “Out” fields are used to control the relays manually.
The "AD.C" & "Rem" fields are used to read the status of these inputs.
Definition of this screen specific keys:
The sixth key is a toggle key, its function toggles from
0/OFF
4000/ON
JUNE 2004
0/OFF
to
4000/ON
Gives 0 pt on all analog outputs and opens all relays contacts.
Gives the full scale (4000 pts) on all analog outputs and closes all relays contacts.
3–39
AF22 MODULE
Environnement S.A
3.3.7
Duplication prohibited
STOP MODE
This screen is used to activate the "Stop mode". To return to measurement mode, it is necessary to
press down the "Sample" key in any screen of Measurement Menu (see Chapter 3.3.2).
Stop mode is used to turn pumping unit off, all the other regulations keep on operation.
3–40
JUNE 2004
Duplication prohibited
AF22 MODULE
3.4
CALIBRATION
3.4.1
OVERVIEW OF CALIBRATION AND CONCEPTS
Environnement S.A
To ensure the accuracy of the measurements performed using the AF22M monitor, the unit must be
regularly checked, calibrated and adjusted, following the quality assurance plan of the user.
– Check of zero and span point:
This operation consists of comparing the monitor response, for zero air and a span point of the
range used, to the gas standards used.
This check is used to measure the monitor drift in time without modifying the adjustment coefficient.
This check can be performed using the internal zero air and span.
Frequency: generally 24 hours in automatic cycle mode.
– 2-point calibration:
This is a procedure for checking and correcting the response of the monitor at zero and at a span
point located at approximately 80 % of the full scale of the measurement range used.
Frequency: monthly, or more frequently if the monitor requires it.
– Span (multi-point calibration):
This involves a complete check up of the monitor performance characteristics (linearity).
Frequency: quarterly, or following out-of-tolerance calibration check results requiring an
intervention on the monitor.
Note about gas generation devices:
For devices providing pressurized gas, it is necessary to provide an excess system to deliver the gas at
atmospheric pressure to the monitor inlet. The materials making up this device should be neutral for the gas
used. When used in an automatic cycle with a cylinder, provide a shut-off solenoid valve that can be remotecontrolled by the monitor (see figure 3.6).
Note about gas cylinders connections:
When the analyzer is equipped with the internal zero filter and/or the internal permeation bench, these
respective inlets are no more available (see figure 2-2), therefore during the calibration procedures the
calibration gas cylinders will be connected to the sample inlet
JUNE 2004
3–41
AF22 MODULE
Environnement S.A
Needle valve
Duplication prohibited
Solenoid valve
Double pressure
regulator
Vent
Gas cylinder
Manifold
External
solenoid valve
control
SPAN INLET
ANALYZER
Figure 3-6 - Pressurized gas connection example
3.4.1.1
Zero air generation
– Check: internal zero filter.
– Calibration: cylinder of reconstituted air or zero air generator. The zero air should be free of any
constituent likely to be measured by the monitor and should not contain more than 0.0005 ppm of
SO2.
3.4.1.2
Span gas generation (SO2)
SO2 calibration gas source standards used for adjustments should be traceable to a National Institute
of Standard & Technology (NIST), Standard reference Material (SRM) or Certified Reference Material
(CRM) in accordance with EPA's Protocol n° 2.
If a permeation device calibration system is used, be sure to allow 24 hours for warm-up.
Zero air used for the dilution of standard gas concentration and for zeroing the analyzer should be the
same and contain less than 0.0005 ppm of SO2.
All fittings, valves, pneumatic lines and other components that may contact span test gas must be
fabricated with highly cleaned inert to SO2 material (i.e. stainless steel).
– Calibration check:
xCylinder of SO2 (1 % accuracy) with a concentration of approximately 80 % of the upper range
limit. The cylinder used should be regularly certified in accordance with the user's quality assurance
plan.
xOptional built in permeation oven with an SO2 wafer device. The concentration generated by the
permeation oven is recorded on the check sheet (only for span check).
xExternal permeation device equipped with an SO2 permeation tube and able to generate a
concentration of approximately 80 % of the upper range limit.
3–42
JUNE 2004
Duplication prohibited
AF22 MODULE
Environnement S.A
– 2-point and multi-point calibration:
xCylinder of SO2 (1 % accuracy) associated to a calibrator allowing the generation of 6
concentrations, including zero, up to 80 % of the upper range limit. The cylinder used should be
regularly certified in accordance with the user's quality assurance plan.
xExternal permeation device equipped with an SO2 permeation tube and able to generate 6
concentrations, including zero, up to 80 % of the upper range limit (refer to section Section10.3 of
Appendix A to 40CFR50).
NOTE :
3.4.1.3
The devices used to generate the span gases should deliver 700 cc/min. minimum and 2500 cc/min.
maximum. All fittings, valves, pneumatic lines and other components that may contact span gases must
be fabricated or coated with cleaned PTFE, glass or stainless steel.
Internal solenoid valves
During multipoint calibration the gas sources will be connected to the sample inlet of the monitor. After
this calibration operation, the gas sources will be connected to their respective inlets and a check will
be made to see that the monitor's response is identical on any inlet used. The various gas inlets can
therefore be used for zero and span point checks and for 2-point calibration of the monitor. Otherwise,
the solenoid valves must be cleaned or replaced.
3.4.2
CHECK OF ZERO AND SPAN POINT
3.4.2.1
Facilities required
– "Zero" air:
a sufficient quality of "zero" air is obtained using the monitor "ZERO AIR" filter (Purafil/activated
vegetable charcoal).
– Span point:
x SO2 gas cylinder, which concentration is below 1 PPM and connected to the monitor "span gas"
inlet.
x Or internal permeation oven with SO2 tube (connected to monitor "span gas" inlet). The
concentration generated by the oven is recorded on the check sheet.
x Or portable calibrator (type VE3M) equipped with an SO2 tube and connected to the monitor
"span gas" inlet.
NOTE : If the analyzer is equipped with an internal permeation oven, the sampling input must be used
to connect the reference cylinder or the portable calibrator.
3.4.2.2
Procedure
– Zero check:
Select the monitor "zero air" inlet using the
Zero
key or appropriate inlet if required and wait for the
measurement to stabilize. The reading should be less than + 2 ppb. If this is not the case, a zero
adjustment is necessary.
– Check of scale point:
Select the gas inlet where the span gas is connected using the
Span
key or the
Sample
key, as per the
above note, and wait for the measurement to stabilize. The result will be compared to the
concentration generated by the device used, taking into account its accuracy.
If the difference between span gas and reading is more than + 10 %, a span adjustment is
necessary.
JUNE 2004
3–43
AF22 MODULE
Environnement S.A
3.4.2.3
Duplication prohibited
Use of automatic cycles
To program the cycles, see section 3.3.3.3 Span Ö Cycles menu.
– Zero cycle:
The "zero" air generator is permanently connected to the monitor "zero air" inlet. The minimum
recommended duration of the zero check is 10 minutes.
– Calibration cycle:
The scale point generator (the internal permeation oven if option is present) is permanently
connected to the monitor "span gas" inlet, it can be the internal permeation oven . The SO2
concentration must be below the full scale of the range used for the measurement. The minimum
recommended duration of the check is 10 minutes.
3.4.3
SPAN ADJUSTMENT
3.4.3.1
Facilities required
– "Zero" air:
See section 3.4.1.1.
– Span point:
See section 3.4.1.2.
3.4.3.2
Procedure
– Zero reference:
x Start by doing a reference cycle (minimum of 600 seconds). Feed zero air to the desired inlet (or
span) and press the
Cycle
key in one of the Measurement screen, then the
Z.Ref
key of the cycle
screen, the cycle duration countdown is displayed in the top right corner of the screen.
– Span adjustment:
x automatic:
Select the measurement range (Configuration Ö Signals/Ranges/Units menu) corresponding to
the concentration (Span Ö Select gas menu) of the standard gas and the inlet used. Press the
Cycle
key, then the
Auto
key of the cycle screen. The analyzer automatically changes its
calibration factor as a function of the concentration programmed. It is advisable that the AUTOCAL lasts 600 seconds, the cycle duration countdown and the concentration of the span gas are
displayed in the top right corner.
NOTE : During the cycle, the SO2 displayed measurement uses the former K span, it is updated
when the AUTO-CAL cycle is achieved.
If the new K span is out of the bounds K(former) r 50 %, the analyzer gives a calibration alarm, in that
case check:
– the calibration pneumatic circuit
– the generated concentration of the gas source used
– the programmed span gas concentration
– the programmed gas inlet
If one of the above points is wrong, correct it and repeat the AUTO-CAL cycle. If nothing is wrong,
proceed to a manual adjustment of the span factor as described after.
3–44
JUNE 2004
AF22 MODULE
Duplication prohibited
Environnement S.A
x manual:
Select the inlet to which the gas is connected by means of the
Sample
or
Span
keys. Wait for the
measurement to stabilize. Program the new K (span) in the Span Ö factors mode menu.
Calculation of new coefficient:
K new = K (former) x
span gas value
read value (without offset)
CAUTION: It is advisable to note K (former) before changing it, since it will be deleted when you enter
K (new).
3.4.3.3
Use the AUTO-CAL automatic cycle
To program this cycle, see paragraph 3.3.3.3 Span Ö Cycles menu. To configure the concentration,
see paragraph 3.3.3.2, Span Ö Select gas menu.
The gas generation system is connected permanently to the monitor "span gas" inlet. The
recommended time for automatic calibration is 600 seconds.
If the span gas meets the requirements described in § 3.4.2.1,this feature could be covered
under US EPA designation.
Flow controler
Flow rate measurement
Dilution air
cylinder or
zero air
generator
Flow controler
Flow rate measurement
Mixing chamber
Doublepressure
reducing
valve
Gas
cylinder
Vent
Manifold
To monitor
Figure 3-7 - Typical calibrator
JUNE 2004
3–45
AF22 MODULE
Environnement S.A
3.4.4
MULTIPOINT CALIBRATION
3.4.4.1
Overview
Duplication prohibited
The minimum device will be made up of a diluter, a certified cylinder of SO2 (1 % accuracy), the
calibration of which can be traced to reference materials from the National Institute of Standards
(NIST) for calibration in the framework of measurements according to the US EPA, and a zero air
generator. Figure 3.7 shows an example of a typical calibrator.
The gases will be applied to the sample inlet of the monitor at atmospheric pressure.
Calibration of the monitor requires generation, in addition to the zero, of 6 gas concentrations (for
example, 15 %, 30 %, 45 %, 60 %, 75 % and 90 % of full scale of range used). The dilution air must
be the same as the zero air.
It is advisable to connect the analog output of the monitor to a strip chart recorder for acquisition and
processing of the data.
3.4.4.2
Equipment Required
– Diluter:
x flow regulators: they must regulate flow rates to the nearest r 1 %.
x flow meters: they must read and record flow rates to the nearest r 2 %.
x mixing chamber: its shape and volume should allow a homogeneous mixture of SO2 and
dilution air.
– Manifold:
The manifold must include at least one outlet for the monitor and one outlet for the excess flow
meter. The outlet must have a sufficient diameter so as not to create any load losses at the inlet of
the monitor. The outlet of the excess flow (at least 20 % of the total flow rate) will be designed so
that the pressure in the manifold is quite near the atmospheric pressure (no overpressure), but so
that the ambient air cannot be back scattered.
– Dilution air (zero air):
Zero air generator or cylinder of reconstituted air free of any contaminants likely to be measured by
monitor and of which the SO2 concentration does not exceed 0.0005 ppm.
– Span gas (SO2):
Certified cylinder of SO2 in air at 1 % of which the concentration will allow generation of 6 ranges
between 15 % and 90 % of the full scale of the measurement range used.
3.4.4.3
Procedure
x Switch on monitor at least 6 hours before making the calibration.
x Configure monitor as follows (advised configuration):
Span ÖCycles menu
REFER. time ................... 0600
AUTO-CAL time .............. 0600
(deactivate all cycles or check that no one of them will be triggered during calibration).
Configuration Ö Measurement mode menu
Resp. time ...................... 01
Configuration Ö Ranges / Units menu
Range 1 - Range 2.....range currently used in measurement
Configuration Ö Offset / Conversion menu
SO2 offset ....................... 10.00 ppb (at least or 5% of full scale of the range used).
3–46
JUNE 2004
AF22 MODULE
Duplication prohibited
Environnement S.A
– Connect analog output (see table 3-1) of monitor to recorder. Use response from recorder to
calibrate device as described below.
– Connect sample inlet of monitor to dilution system manifold.
– Adjust dilution system so that it generates a total zero air flow rate that is more than 20 % greater
than the sample flow rate of the monitor.
– Generate zero air. Wait for reading to stabilize (recommended duration: 600 sec). Note value of
ZSO2, expressed in percentage of maximum recorder response (for example, 400 mV / 10 V = 4 %).
– Follow steps in section 3.4.3.2. ("span point correction" section) to adjust calibration coefficient.
Use formula below to calculate value of span generated:
[SO2] generated
[SO2] Cylinder u FSO2
FD FSO2
where:
[SO2]Generated is the concentration of SO2 in the gas generated at the outlet of the manifold,
[SO2]Cylinder is the concentration of SO2 in the certified cylinder,
FSO2 is the flow rate of SO2 in NI/min,
FD is the dilution air flow rate in NI/min.
Use formula below to calculate the reading value in ppm from the value in % full scale to be taken
into account:
[SO2] read
§ Srecord ZSO2 ·
¨
¸ u ECH
©
¹
100
where:
S record is the value noted on the recorder, expressed in percentage of the full scale of the
recorder,
ECH is the full scale of the monitor,
ZSO2 was measured previously.
– Then generate the other five SO2 concentrations between 15 and 90 % of the full scale, varying the
flow rates FSO2 and/or FD.
– Plot the [SO2]read values as a function of the [SO2]generated values, including the zero air point.
Check linearity.
– Plot or calculate, using the formula for least error squares below, the straight line that will constitute
the straight calibration line for the monitor.
[SO2] read
a x [SO2] generated b
where:
a ; is the linear regression coefficient (slope), and is calculated as follows:
a
n x ¦ [SO2] generated x [SO2] read ¦ [SO2] generated x ¦ [SO] read
n x ¦ [SO2] 2generated ¦ [SO2] generated
2
b ; is the constant linear regression term (intercept) ; it is calculated as follows:
b
JUNE 2004
¦ [SO2] read a x ¦ [SO2] generated
n
n is the number of items of data.
3–47
AF22 MODULE
Environnement S.A
3.4.5
Duplication prohibited
INTERNAL PERMEATION OVEN (OPTIONAL)
The scale point check can be performed using a permeation oven equipped with a SO2 source.
– Principle:
The SO2 used is maintained in a state of liquid/vapor phase equilibrium in a closed cylindrical tube
fitted with a polymer diaphragm. Due to the differential partial pressure of the gas on either side of
the diaphragm and the diaphragm "permeability", gas is diffused toward the outside of the tube.
The weight of the gas diffused by unit of time, called "permeation rate", depends on several
parameters: type of gas, thickness, surface and nature of diaphragm, partial gas pressures on
either side of diaphragm, temperature.
– Setup:
The permeation tube is placed in a block which is thermostat-controlled to 50 °C (± 0,1 °C) and
continuously ventilated with a scavenging gas whose flow is regulated to about 45 l/h. The rinsing
gas is filtered by an active charcoal filter located at the rear of the analyzer and directly connected
to the inlet of the permeation oven. The outlet of the permeation oven (PTFE plug) is connected to
the "span gas" inlet of the analyzer.
When this inlet is selected, the monitor samples a part of the generated gas and thus measures the
concentration of this gas.
– Procedure:
x The internal oven is selected using the "span" key, or automatically on a calibration cycle. Once
stabilized, the SO2 measured by the monitor must be compared to the concentration recorded
on the check sheet.
x Since SO2 permeation is very sensitive to the quality of the rinsing air, the condition of the
activated vegetable charcoal filter should be checked periodically.
x The oven temperature is set using the Tests Ö Mux signals function on channel No. 14 (see
Table 3.3).
x The system operating temperature range is 10 to 35 °C (ambient temperature).
x When setting into service or following an extended shutdown period, the stabilization time is
about 24 hours.
x The permeation source supplied with the monitor has an autonomy of about 16 months.
If the monitor is not used, the source should be removed from the oven, and placed in its
original package with the desiccant and absorbent bags, then stored in a cool, ventilated
location.
3–48
JUNE 2004
Duplication prohibited
AF22 MODULE
Environnement S.A
– To maintain the best precision, the following characteristics should be checked regularly:
x flow rate of oven diluent air,
x permeation tube rate (theoretically constant).
A differential weighing using an accurate balance should be performed to the 10 th, or better to the
100 th of mg.
Calculation of permeation (P) and (CG) concentration rates generated by oven:
m0 = Initial weight of tube (ng)
m1 = Final weight of tube (ng)
m0 m1 = Weight of SO2 diffused (ng)
t = Time between two weighings (min)
P = (m0 m1) / t = Tube permeation rate (ng/min.)
F = Real total air flow rate passing through the oven (CC/min.)
Km = Molar coefficient, for SO2 = 0.382
CG = Km x P / F = Concentration of span gas generated (ppm)
NOTE :
These calculations must be repeated each time a permeation tube is changed, or, more simply, the
concentration generated by the new tube can be determined by reading the concentration on the oven
just after calibration of the analyzer (see section on calibration).
JUNE 2004
3–49
AF22 MODULE
Environnement S.A
Duplication prohibited
Page intentionally left blank
3–50
JUNE 2004
AF22 MODULE
Duplication prohibited
Environnement S.A
CHAPTER 4
PREVENTIVE MAINTENANCE
4.1
SAFETY INSTRUCTIONS
4–2
4.2
MAINTENANCE CALENDAR
4–3
4.3
MAINTENANCE OPERATION SHEETS
4–4
4.4
AF22M MAINTENANCE KIT FOR 1 YEAR
JULY 2002
4–11
4–1
AF22 MODULE
Environnement S.A
4
PREVENTIVE MAINTENANCE
4.1
SAFETY INSTRUCTIONS
Duplication prohibited
The safety instructions must be observed at any time by the user.
Whenever possible, cut off the power supplies when performing any work on the monitor.
Only qualified personnel should intervene on the monitor.
As concerns safety, the manufacturer shall not be responsible for any consequences resulting from:
use of the monitor by non-qualified people,
use of the monitor under conditions other than those specified in this document,
modification of the monitor by the user,
no-maintenance of the monitor.
A systematic periodic inspection is required.
4–2
JULY 2002
AF22 MODULE
Duplication prohibited
4.2
Environnement S.A
MAINTENANCE CALENDAR
By its design, the AF22M requires very limited maintenance. However, to ensure the monitor's
performance characteristics over time, the unit must be serviced regularly. The periodicities indicated
below are given as an example and can vary according to operating conditions.
Operation
– Sample inlet PTFE filter
– Check of electrical parameters
– Check of zero and calibration
– Replacement of activated charcoal (internal zero filter) and
kicker zero filter cartridge
– Inspection of pump valves and diaphragms
– Replacement of UV lamp
Periodicity
Sheet No.
15 days
4.3.1
1 month
4.3.2
1 month
4.3.3
One year
4.3.4
Six months
4.3.5
2 years
4.3.6
Annual check
The monitor must be returned to laboratory for complete cleaning (measurement chambers,
restrictors, flow circuit, etc.) and check of all metrological parameters.
JULY 2002
4–3
AF22 MODULE
Environnement S.A
4.3
Duplication prohibited
MAINTENANCE OPERATION SHEETS
MAINTENANCE SHEET
MONITOR Serial No.:
Scope: Replacement of filters
OPERATION SHEET: 4.3.1
PAGE: 1/1
Periodicity: 15 days
Sample inlet PTFE filter: "MITEX" Teflon filter - 5 μm porosity - 47 mm dia.
Ref.: F05-11-842
Date
a) Unlock the filter cover (2) by rotating the locking spring (1)
b) Remove the dirty filter (3)
c) Place the new filter (4) by slipping it slightly from its paper protection(5) onto the
filter holder (6).
2
1
3
5
4
6
Figure 4-1 – Inlet dust filter
Tools required
x
none
4–4
JULY 2002
AF22 MODULE
Duplication prohibited
Environnement S.A
MAINTENANCE SHEET
MONITOR Serial No.:
Scope: Check of electrical parameters
OPERATION SHEET: 4.3.2
PAGE: 1/1
Periodicity: 1 month
– Electrical values are checked using the Tests Ö Mux signals (see Chapter 3, paragraph 3.3.6.2).
– Compare the voltage values at each input of the multiplexer to the values given for the same inputs on the
test sheet provided with the monitor.
Mux Reading
Dates
2
6
7
8
9
10
11
12
14
– Tools required
x None.
JULY 2002
4–5
16
AF22 MODULE
Environnement S.A
Duplication prohibited
MAINTENANCE SHEET
MONITOR Serial No.:
Scope: Checks of zero and calibration
OPERATION SHEET: 4.3.3
PAGE: 1/1
Periodicity: 1 month
ZERO
Span
reading
reading
Dates
– Check of zero (see section 3.4.2)
Zero checking operations can be carried out:
x Either using a pressure-packed, synthetic air cylinder
guaranteed to be free of any trace of SO2, and injected
in the analyzer sample inlet at atmospheric pressure.
x or using a zero air provided by the pure-air generator,
for example of the AADCO type, and injected under
the same conditions as the previous example.
x or using the internal zero filter, by actuating the
Zero
key, which controls the internal solenoid valve.
– Manual calibration of monitor (see section 3.4.3).
It is highly advisable to carry out this operation on site,
under regular operating conditions. Once the calibration is
made, the connections of the analyzer must not be
handled.
This operation requires the use of a SO2 in synthetic air
gas cylinder for calibration (this cylinder should be fitted
with a double pressure relief valve) or the use of a
calibration device (portable calibrator).
The gas used for calibration should be delivered at
atmospheric pressure (for example, through a manifold at
the "span gas" inlet of the analyzer).
– Tools required
x None
4–6
JULY 2002
Duplication prohibited
AF22 MODULE
Environnement S.A
MAINTENANCE SHEET
MONITOR Serial No.:
Scope: Replacement activated charcoal (internal zero
filter) and kicker zero filter cartridge.
OPERATION SHEET: 4.3.4
PAGE: 1/1
Periodicity: Six months
Date
– Disconnect and remove used cartridge.
– Connect the tube to the fitting provided for this purpose on the Zero solenoid valve.
– Position cartridge.
1
2
Figure 4-2 – Zero filter cartridges position
NOTE : The built-in zero filter (1) lifetime is 1 year.
The hydrocarbon elimination device scavenging air zero filter (2) is 6 months
(double flow rate than filter (1) and continuous working)
– Tools required
x None
JULY 2002
4–7
AF22 MODULE
Environnement S.A
Duplication prohibited
MAINTENANCE SHEET
MONITOR serial No.:
OPERATION SHEET: 4.3.5
Scope: Check of pump valves and diaphragm
PAGE: 1/1
– Check condition of diaphragm and pump valves. Replace them if necessary.
Frequency: Yearly
Type of
operation
Date
Replacement of diaphragm and valves
– The diaphragms and valves are parts that wear out; these parts are replaced
very easily.
Replacement of diaphragm and pump valves
– Power instrument off, unplug the power cord
a) Disconnect supply connector.
b) Disconnect gas flow connections.
c) Loose the 3 screws (1) fastening pump assembly to frame in order to release
the 3 pump holder slots.
d) Unscrew the 2 screws (3) fastening pump on its holder.
e) Draw a permanent mark (13) on the pump body (8) for reassembling
purpose.
f) Remove the nut (5) fastening pump body (8) on the vibrating blade (6).
g) Unscrew the 2 screws (7) fastening pump body to pump housing (9).
h) Remove pump body.
i) Separate the various parts..
j) Clean diaphragm (10), the two valves (11) and the gasket (12).
k) Dust off or clean the different parts with alcohol (dismount valves if only
necessary)
l) Check diaphragm, change it if necessary.
m) Reassemble, repeating steps above in reverse order.
– Tools required
x Dia. 4 mm cross-tip screwdriver,
x Dia. 5 mm flat screwdriver,
x Combination pliers
x Alcohol solution
Type of operation:
D: Replacement of diaphragm
4–8
V: Replacement of valves
JULY 2002
AF22 MODULE
Duplication prohibited
Environnement S.A
MAINTENANCE SHEET
MONITOR serial No.:
OPERATION SHEET: 4.3.5
Scope: Check of pump valves and diaphragm
PAGE: 2/2
Frequency: Yearly
1
10
2
3
11
4
12
5
6
13
7
8
9
Figure 4-3 – Exploded view of the pump.
JULY 2002
4–9
AF22 MODULE
Environnement S.A
Duplication prohibited
MAINTENANCE SHEET
MONITOR Serial No.:
Scope: Replacement of UV lamp
OPERATION SHEET: 4.3.6
PAGE: 1/1
Periodicity: 2 years
WARNING: Do not touch the glass part of the lamp.
a) Switch off the monitor.
b) Disconnect the 2 supply wires of the lamp, loosening them from the UV supply board.
c) Unscrew the 2 screws (1) to remove the lamp / shutter assembly.
d) Loose off 3 mm the 2 screws (2) to release the UV lamp.
e) Replace the UV lamp.
f) Align mechanically the new lamp, taking into account the MAX OUTPUT PORT mark (3). This window must
face the hole (4).
g) Remount the lamp assembly.
h) Select the Tests Ö MUX signals screen:
x Check the voltage and current of the UV lamp. I-UV supply: 350-450 mV adjusted on UV lamp supply
board
x Adjust the position of the lamp by rotating and translating it so as to have the maximum UV signal
(channel
9
of
MUX)
voltage.
Adjust to approximately 4300 mV.
i) Tighten the 2 screws (2) when maximum signal is obtained to fix the UV lamp.
j) After stabilization of the UV power, run a reference cycle, and then calibrate the analyzer.
1
3
4
2
2
1
Figure 4-5 – Lamp / shutter assembly (rear face)
Figure 4-4 – Lamp / shutter assembly (front face)
– Tools required
x 3.5 x 75 mm - 2.5 x 50 mm screwdrivers
4–10
JULY 2002
Duplication prohibited
4.4
AF22 MODULE
Environnement S.A
AF22M MAINTENANCE KIT FOR 1 YEAR
Kit reference : AF22-K
This kit consists in:
1 box of 25 filtering membranes for sample inlet particulate filter
SAV-K-000042-A
3 filtering cartridges
F05-0128-1
1 maintenance kit for internal pump
V02-K-0041-A
JULY 2002
4–11
AF22 MODULE
Environnement S.A
Duplication prohibited
Page intentionally left blank
4–12
JULY 2002
AF22 MODULE
Duplication prohibited
Environnement S.A
CHAPTER 5
CORRECTIVE MAINTENANCE
Table 5–1– List of faults and corrective actions
Table 5–2 – AF22M Module board configuration
Table 5–3 – Board RS4i Configuration
Table 5–4 – Keyboard Interface configuration
Table 5–5 – UV lamp board configuration
Table 5–6 – Configuration of UV lamp supply board
Table 5–7 – Flow rate control board configuration and setting
5–4
5–9
5–10
5–11
5–12
5–13
5–14
Figure 5-1 – AF22M Module board
Figure 5-2 – Card RS4i Configuration
Figure 5-3 – Keyboard Interface board
Figure 5-4 – UV lamp board
Figure 5–5 – UV lamp supply board
Figure 5-6 – Flow rate control board
5–8
5–10
5–11
5–12
5–13
5–14
MAY 2007
5–1
AF22 MODULE
Environnement S.A
Duplication prohibited
Page intentionally left blank
5–2
JUNE 2004
Duplication prohibited
5.
AF22 MODULE
Environnement S.A
CORRECTIVE MAINTENANCE
Corrective maintenance of the monitor should only be performed by qualified people using the
information provided in this document.
The monitor automatically and continuously self-tests its main components. Any malfunction detected
is indicated by a clear message on the display and a buzzer.
Table 5.1 summarizes the main faults indicated by the unit with corresponding corrective possible
actions.
In case of operating fault, the ALARM message blinks in the top right corner.
To check which operating fault is present select the menu Measurement Ö Alarms display.
JUNE 2004
5–3
AF22 MODULE
Environnement S.A
Duplication prohibited
Table 5-1– List of faults and corrective actions
ALARM MESSAGE
CHOPPER FAULT
OVERRANGE
Optical T°C FAULT
Flow rate Fault
CAUSE
POSSIBLE ACTION
– The shutter does not cut off
the UV beam.
– Check that connector J9 is indeed inserted.
– Check that the shutter is not blocked by its
supply cable.
– Check the solenoid control.
x To do this: start a Zero Ref. cycle.
x Put a multimeter between the two
terminals of the connector. For the first
40 seconds of the cycle, the multimeter
should give a voltage reading of 24 V.
x If it does not, disconnect J9. Check the
electrical continuity of the coil of the
solenoid (between end terminals of J1).
Replace
solenoid
if
necessary.
Otherwise, check mechanical positioning
of the shutter. If you still do not locate the
problem, contact us.
x If this is not the case: check that the 5 V
is available in the control transistor
emitter, located on the motherboard,
throughout the duration of the control. If
the voltage is indeed available, check the
availability of the D.C. power supply. If
you still do not locate the problem,
contact
us.
– The measurement value
exceeds the range 2.
– Change ranges.
– The temperature of the optical – Check that connector J17 is correctly
bench does not fall within the
inserted.
boundary values for correct
Run a warm up of the monitor by stopping it
operation
for a few seconds, then switching it back on.
(38 °C < T < 48 °C)
Check the availability of the 24 V voltage at
both end terminals on J17.
– The temperature probe
is defective.
– Contact us.
– The Module board is
defective.
– Replace board or contact us.
– Incorrect flow rate in
measurement chamber; the
pump is not working.
– Check connection. If everything is OK,
overhaul pump completely.
– Leak in fluid circuit.
– Check all connections.
– Place a flow meter at the outlet of the
monitor. Plug the sample or calibration inlet,
according to the case. The ball in the flow
meter should drop to zero. If it does not,
there is a leak.
5–4
JUNE 2004
Duplication prohibited
ALARM MESSAGE
Calibration fault
UV Power FAULT
AF22 MODULE
CAUSE
Environnement S.A
POSSIBLE ACTION
– Span gas source
– Check the availability of the span gas at
atmospheric pressure on the selected inlet.
– Following a manual reset of
the span K factor
– Run a manual calibration.
– The control voltage of the UV
power does not fall in the
boundary values for correct
operation 2,500 mV <
UV voltage < 4,500 mV.
– Check that connector J7 is correctly
inserted.
Check the stability of the regulated voltage
between the right-hand end point of
connector J7 and the ground. If one of these
voltages is off, contact us.
– The voltage function of lamp
current does not fall in the
boundary values 80 < UV cur
< 440
Set voltage between 350 mV and 450 mV
using potentiometer P (MUX 12) I. UV.
x move the lamp to adjust maximum UV
energy.
If it is still not possible:
x increase the UV lamp supply current
using SPAN Ö E2Pot screen, I (mA)
lamp function.
If it is still impossible to make the setting, or
if the voltage is regulated but the failure
persists, contact us.
– After stabilization of the UV power, run a
Zero Ref. cycle, then calibrate the analyzer.
UV SOURCE FAULT
Pressure Fault
Permeation T° Fault
ZERO filter life time
MAY 2007
– The UV signal does not fall in
the boundary values for
correct operation 1000 mV <
UV signal < 9800 mV.
– The signal given by the
pressure sensor (optional)
doesn't fall in the limits 500 <
Pressure < 1050
– Refer to possible actions described for UV
power fault
– Adjust the SPAN Ö E2Pot, UV signal
function.
– Check that sample is indeed at atmospheric
pressure.
– Check that there is no clogging.
– Disconnect the tube on pressure sensor.
Check that measurement corresponds to the
ambient barometric pressure.
– Ambient working temperature
out of bounds
– Permeation oven temperature – +15V permeation oven pcb's fault.
regulation fault
– The zero filter life time down
– Replace the filter (see maintenance sheet
counter has reached 0.
4.3.4)
– Reset the zero filter life time down counter
(ConfigurationÖMeasurement mode
screen).
5–5
AF22 MODULE
Environnement S.A
Duplication prohibited
Table 5-1– List of faults and corrective actions
SYMPTOM
(no fault indication)
Monitor does not react when
switched on.
POSSIBLE CAUSES
– Mains fault
– Check for presence of mains power supply.
– Faulty mains cable
– Test continuity of mains cable.
– Connector
plugged in
– Check fuse in connector assembly.
– Mains
Monitor remains in preheating
status
ACTIONS
incorrectly
fuse
blown
– Module board faulty.
– Check that display flashes.
– Reset circuit blocked
Micro 5 V faulty.
– If not flashing, check if the microprocessor
board is in place.
– Replace if necessary.
– If flashing, wait for 15 min. and possible fault
indication.
Switching power supply is protected from short circuits. In such case, it is
necessary to disconnect / re-connect the power cord to reset
STOP
5–6
JUNE 2004
Duplication prohibited
AF22 MODULE
Environnement S.A
Page intentionally left blank
JUNE 2004
5–7
AF22 MODULE
Environnement S.A
Duplication prohibited
J1
J2
PT4
J3
PT5
PT3
PT2
J4
J5
PT4
PT1
J6
J35
ST3
J7
ST1
J8
PT16
PT17
PT7
J9
ST2
J10
PT9
SW4
J11
SW3
J12
Pt 14
J13
J14
J19
J15
PT11
J16
J17
J18
PT8
Pt 13
Pt 12
Figure 5-1 – AF22M Module board
5–8
JUNE 2004
AF22 MODULE
Duplication prohibited
Environnement S.A
Table 5-2 – AF22M Module board configuration
Type of signal
Test points.
Connection
Connector
PT1
Flow rate
J1
PT2
Chamber pressure
J2
O2 option
UV detector
PT3
PM signal
J3
PM preamplifier
PT4
Gas temperature
J4
Gas temperature
PT5
A/D RUN
J5
Chamber temperature
PT6
A/D STATUS
J6
Flow rate sensor
PT7
Permeation bench
J7
UV lamp supply
PT8
0 V (GND)
J8
Permeation bench
PT9
H2S converter
J9
Chopper
PT10
MUX output
J10
H2S/TRS converter
PT11
+24VCC
J11
Fan
PT12
+5VCC
J12
Zero SV
PT13
+15 VCC
J13
Span SV
PT14
-15VCC
J14
H2S SV
PT15
Not used
J15
Pump 2 (permeation)
PT16
I off+
J16
Flow rate
PT17
I off -
J17
Chamber heating
J18
+24VCC supply
J19
+24VCC on/off switch
J20
Estel board I2C bus
J21
Synchro bus
J22
RAM extention option
J23
RS4i board I2C bus
J24
Not used
J25
Not used
J26
Not used
J27
Not used
Nature of operation
Jumpers
SW1
EPROM
Configuration
SW3
Power supply
SW4
Internal clock
ST1
ST2
Watch dog
JUNE 2004
1
2
3
Selects 27C20 (2 Mbytes)
1
2
3
Selects 27C40 (4 Mbytes)
1 2 3
5V I2C Bus
J28
Not used
1 2 3
24V I2C Bus
J29
Not used
1 2 3
=CLK/2
J30
Not used
1 2 3
= CLK (default position)
J31
Not used
(default position)
J32
Not used
Resets microprocessor
J33
Not used
Inactive
J34
Stand by LED option
Active (default position)
J35
Maintenance switch option
5–9
AF22 MODULE
Environnement S.A
Duplication prohibited
Table 5-3 – Board RS4i Configuration
Jumpers
references
SW1, SW2
SW3
Nature of operation
Channel 1 on RS422 standard
Channel 1 on RS232 standard
Load RX bus RS422 active
ST1
Load RX bus RS422 inactive
Load TX bus RS422 active
ST2
Load TX bus RS422 inactive
Not used
ST3
NOTE :
Symbols
Channel 2 is on the RS232 standard.
ST3
SW2
SW3
SW1
ST2
ST1
Figure 5-2 – Card RS4i Configuration
5–10
JUNE 2004
AF22 MODULE
Duplication prohibited
Environnement S.A
Table 5-4 – Keyboard Interface configuration
Jumpers
references
Symbols
Nature of operation
Keyboard inhibited
ST1
Keyboard active
P1
LCD contrast adjustment potentiometer
P1
ST1
Figure 5-3 – Keyboard Interface board
JUNE 2004
5–11
AF22 MODULE
Environnement S.A
Duplication prohibited
Table 5-5 – UV lamp board configuration
References
P1
Nature of operation
Lamp current pre-adjustment
PT1
GND
PT2
LM2596 output (SV)
PT3
Frequency (25 kHz)
PT4
V. f(I)
PT5
I lamp adjust
PT6
I lamp
UV lamp supply
PT4
PT5
P1
PT6
PT2
PT1
PT3
Figure 5-4 – UV lamp board
5–12
JUNE 2004
Duplication prohibited
AF22 MODULE
Environnement S.A
Table 5–6 – Configuration of UV lamp supply board
Jumpers
SW1
Nature of operations
3
2
1
AF22M
Measure lamp (2-3 are linked)
ST1
ON
ST2
OFF
Figure 5–5 – UV lamp supply board
MAY 2007
5–13
AF22 MODULE
Environnement S.A
Duplication prohibited
Table 5-6 – Flow rate control board configuration and setting
Designation
Nature of operations
Ground referenced to Module
Board
ST1
Direct ground (default position)
P1
Flow rate adjustment
P2
Pump frequency adjustment
PT0
Ground
PT1
Flow rate signal
PT2
Pump supply voltage
P2
PT2
PT2
ST1
P1
PT0
Figure 5-5 – Flow rate control board
5–14
JUNE 2004
Duplication prohibited
AF22M
Environnement S.A
6
CHAPTER 6
APPENDIX
ESTEL board
SOREL board
JUNE 2004
6–1
AF22 MODULE
Environnement S.A
Duplication prohibited
Page intentionally left blank
6–2
MAY 2007
ESTEL Board
INPUTS / OUTPUTS BOARD
OPTION OF 2M
ANALYZERS
- June 2009 -
WARNING
Information contained in this document are likely to be modified without notice.
The designer reserves the right to modify the equipment without improving this document,
therefore, information of this document does not represent a commitment under ENVIRONNEMENT S.A.
ENVIRONNEMENT S.A. all right reserved.
111 bd Robespierre, 78300 POISSY - -TEL. 33(0)-1.39.22.38.00 – FAX 33(0)-1.39 65.38.08
http://www.environnement-sa.com
ESTEL Board
Environnement S.A
Duplication prohibited
ESTEL BOARD
1.1
FUNCTION AND USE
3
1.2
TECHNICAL CHARACTERISTICS
3
1.3
CONFIGURATION
4
1.4
PROGRAMMATION
8
1.5
1.6
1.4.1
ESTEL CARD(S) Ö Analog output
1.4.2
ESTEL CARD(S) Ö Analog input
11
1.4.3
ESTEL CARD(S) Ö Relay
12
1.4.4
ESTEL CARD(S) Ö Remote controls
13
9
INSTALLATION AND REPLACEMENT OF ESTEL BOARD
14
1.5.1
Switch off the analyzer
14
1.5.2
Unplug the mains cable
14
1.5.3
Put off the cover
14
1.5.4
Dismount ESTEL board
15
1.5.5
Unrivet the back plate (6) of rear panel of analyzer
15
1.5.6
Installation of board inside the analyzer
16
OPTION OF EXTERNAL CONNECTION
17
Figure 1 - ESTEL board_index A
Figure 2 – ESTEL board-index B
Figure 3 – Option of external connection P10-1337-A
Figure 4 – Option of external connection + 4 insulated outputs P10-1338-A
5
6
18
18
Table 1 - Configuration of ESTEL board_index A
Table 2 - Configuration of ESTEL board_index B
5
6
Up-to-date:
Pages
Up-to-date
Pages
Up-to-date
1
05-2004
9
06-2009
17
06-2009
2
05-2004
10
06-2009
18
06-2009
3
05-2004
11
06-2009
4
05-2004
12
06-2009
5
05-2004
13
06-2009
6
05-2004
14
06-2009
7
05-2004
15
06-2009
8
06-2009
16
06-2009
2
Pages
Up-to-date
JUNE 2009
ESTEL Board
Duplication prohibited
1.
Environnement S.A
ESTEL BOARD
ESTEL board is a universal board of logic and analog inputs/outputs for the 2M series analyzers. It is
optional: it is possible to install up to 2 ESTEL boards in an analyzer.
1.1
FUNCTION AND USE
The ESTEL board has 4 functions:
x
4 Analog inputs,
x
4 Analog outputs,
x
6 Relays,
x
4 Remote controls.
The ESTEL board enables dialog with the measurement module and relieves it of the Inputs/Outputs
functions. It enables remote control and/or remote signalling of certain functions as: "measurement",
"zero", "calibration", "alarm".
The analog inputs are used to connect independent monitors in order to follow-up, for example,
weather parameters.
The analog outputs enable to send numeric parameters (gas concentration to be analyzed, MUX
channels) to analog independent peripherals in order, for example, to store and process several data
of several months.
Equipped with an ESTEL board, the analyzer can work as an autonomous unit of analysis.
1.2
TECHNICAL CHARACTERISTICS
Management by specialized micro controller:
x
4 analog inputs of 12 bits, 0-2,5 volts full scale,
x
4 not-insulated analog outputs, configurable into: 0-1 volts, 0-10 volts, 0-20 mA, 4-20 mA
(maximum load of 1000 Ohm).
x
4 insulated by optocoupler logic inputs,
x
6 potential-free contacts for remote signalling,
x
only one power supply of 8 to 24 volts,
x
i2C communication visualization using a LED.
Electric connection:
x
4-point connector for link with Module boards of 2M series,
x
Inputs / Outputs centralized on only one female connector SUB D 37 points. This connector
is screwed on the rear panel of the analyzer.
x
Option of external connection, see paragraph 1.6
Voltage and current on relays:
x
Maximum voltage by relay contact : 50 volts
x
Maximum current by relay contact : 1 Ampere at 24 V D.C. (resistive load)
Remote controls:
x
MAY 2004
By dry contact between (1-4) Remote control and ground Remote control.
3
ESTEL Board
Environnement S.A
1.3
CONFIGURATION
PIN N°
CONNECTION
1+
20 GND
Analog output 1
2+
21 GND
Analog output 2
3+
22 GND
Analog output 3
4+
23 GND
Analog output 4
5+
24 GND
6+
25 GND
7+
26 GND
8+
27 GND
Duplication prohibited
PIN N°
CONNECTION
14-33
Relay contact 1
13-32
Relay contact 2
12-31
Relay contact 3
11-30
Relay contact 4
10-29
Relay contact 5
9-28
Relay contact 6
15 +
34 GNDI
Remote control 1
16 +
35 GNDI
Remote control 2
17 +
36 GNDI
Remote control 3
18 +
37 GNDI
Remote control 4
19
5 VCC or + 24 VCC
Analog input 1
Analog input 2
Analog input 3
Analog input 4
(*) according to SW5 jumper position
GND: ground
GNDI: insulated ground
4
MAY 2004
ESTEL Board
Duplication prohibited
Environnement S.A
Table 1 - Configuration of ESTEL board_index A
Jumpers
mark
Symbols
Nature of operations
ESTEL selection, board N° 1
ESTEL selection, board N° 2
ST1, ST2, ST8
ESTEL selection, board N° 3
ESTEL selection, board N° 4
ST3
0 V to ground (default)
Floating 0 V
0-1 V, idem for the 4 DAC.
DAC1
DAC2
DAC3
DAC4
0-10 V, idem for the 4 DAC.
0-20 mA, idem for the 4 DAC.
4-20 mA, idem for the 4 DAC.
P1, P2, P3, P4
4 mA adjustment in 4-20 mA mode
Figure 1 - ESTEL board_index A
MAY 2004
5
ESTEL Board
Environnement S.A
Duplication prohibited
Table 2 - Configuration of ESTEL board_index B
Jumpers mark
Symbols
Nature of operations
ESTEL selection, if 1 board
ESTEL selection, if 2 boards
ST7, ST6, ST5
ESTEL selection , if 3 boards
ESTEL selection, if 4 boards
ST8
0 V to ground (default)
Floating 0V
DAC 1
0-1 V (or optional 2,5 V and 10 V)
idem for the 4 DAC
DAC 2
0-10 V, idem for the 4 DAC
DAC 3
0-20 mA, idem for the 4 DAC
DAC 4
4-20 mA, idem for the 4 DAC
SW5
Output 5 V
Output 24 V
on pin 19
Figure 2 – ESTEL board-index B
6
MAY 2004
ESTEL Board
Duplication prohibited
Environnement S.A
Specific configuration of output 0-5 volts instead of 0-10 volts
There are 4 possible configurations for 0-5 volts output:
x
Board configured into 0-10 volts with addition of a by-2 divider bridge :
The user (customer) carries out himself the operation at input of his acquisition system.
Operating mode:
Connect each analog output, previously configured into 0-10 volts, to ground through 2 resistances
of equal value within 500 and 1000 ohms.
Take off the divided-by-2 signal at terminals of the resistance that is connected to the ground.
0-10 V
R
0-5 V
R
Ground
x
R = 500 ohms
Board configured into 0-10 Volts with adjustment of half gain :
In menu Tests Ö ESTEL boards, adjust the A and B coefficients of each channel in order to obtain
0-5V at analog output for 0-4000 pts resolution of analog-to-digital converter.
x
Board configured into 0-20 MA
The user (customer) carries out himself the operation at output of his acquisition system.
Operating mode:
Connect each analog output, previously configured into 0-20 mA, to ground using a 250 0hms
resistance, tolerance 1 %.
Voltage, thus generated, is equal to UmV = 250 x ImA, that is to say 5 V for I = 20 mA.
Note: place resistance the nearest possible of receiver equipment.
x
Modification of gain resistance on ESTEL board
We carry it out if the user (customer) does not accept the other solutions.
MAY 2004
7
ESTEL Board
Environnement S.A
1.4
Duplication prohibited
PROGRAMMATION
The ESTEL board programmation is carried out from the « ESTEL board » menu of the « Carte(s)
I2C » screen.
This menu enables to visualize the effective communications of the various modules and to configure
the various ESTEL boards.
The analyzer automatically detects the presence of one or several ESTEL boards and offers menus
enabling the user to adjust and configure each board.
8
JUNE 2009
ESTEL Board
Duplication prohibited
1.4.1
Environnement S.A
ESTEL CARD(S) Ö Analog output
To access the various screens of the ESTEL board, select the current function and choose the wanted
function using the [ K ], [ L ] keys.
« Analog output » Function
This screen enables to assign the parameters to the analog outputs for the ESTEL board whose n°
indicated in the field « No » is highlighted. These parameters are:
x
Concentration of the analyzed gases,
x
Auxiliary channels (multiplexer),
x
Analog inlets.
REMINDER : On an ESTEL board, analog outputs can be configured into: 0–1 Volt, 0–10 Volt, 0-20
mA, 4–20 mA.
Four ranges are available and correspond to the full scale of the analog output, units are those of the
parameters displayed in the « Signal » column.
When signal value is higher than the full scale of the current range, the analyzer switches to the next
higher range. It switches again to the lower range when measurement again passes under 85%.
JUNE 2009
9
ESTEL Board
Environnement S.A
Duplication prohibited
When the user assigns several measurement ranges to only one analog output, he can change the
metrological resolution as the here-below curve shows it.
OUTPUT SIGNAL
1V
0
1
10
2
100
3
1000
4
10000
MEASUREMENT RANGES (PPM)
To avoid the ranges switching, the user must assign the same value to the 4 ranges of the parameter
he will send to an analog output.
The Ax+B calibration curve is used to adjust the mV signal of the taken-into-account analog output.
The « Test » column is used to test the 5 analog outputs and to adjust the points number.
For a range 1 :
– 0 point (lower scale of output) @ 0 volt obtained at output,
– 4000 points (higher scale of output) @ 1 volt obtained at output.
F6 key [ Points 4000 ] enables to force the full scale on the whole analog outputs.
10
JUNE 2009
Duplication prohibited
1.4.2
ESTEL Board
Environnement S.A
ESTEL CARD(S) Ö Analog input
Each ESTEL board has 4 analog inputs : this screen is used to program characteritics of these analog
inputs.
JUNE 2009
x
"Name" fields are used to enter a name of 8 alphanumeric digits.
x
"Unit" fields are used to select the unit by : none, ppt, ppb, ppm, μg/m3, mg/m3, gr/m3, μg/Nm3,
mg/Nm3, gr/Nm3, μg/Sm3, mg/Sm3, gr/Sm3, %, μgr, mgr, gr, mV, U, °C, °K, hPa, mb, b,l, Nl, Sl,
m3, l/min, NI/min, Sl/min, m3/h, Nm3/h, Sm3/h, m/s ou km/h, in the toggle menu.
x
The "Ax + B" fields enable to adjust the calibration curve of each parameter.
11
ESTEL Board
Environnement S.A
1.4.3
Duplication prohibited
ESTEL CARD(S) Ö Relay
"Relays" fields are used to control relays according to the following conditions:
Disable
Ö Relay not assigned
General alarm
Ö Any operating fault triggers the relay
Range over-range
Ö Scale 2 over range triggers the relay
Flow rate
Ö Abnormal flow rate triggers the relays
Temperature
Ö Abnormal temperature in the analyzer triggers the relay
Pressure
Ö Barometric pressure in chamber
Zero Air
Ö On Zero, relay is triggered
Span
Ö On Span, relay is triggered
Zero-Ref
Ö On Zero-Ref, relay is triggered
Auto Span
Ö On Auto-Span, relay is triggered
Warm-Up
Ö On Warm-up, relay is triggered
Stop mode
Ö In Stop mode, relay is triggered
Alarm control
Ö Control detection during threshold over range, relay is triggered.
Alarm or Control
Ö Relay triggered
Module alarm
Ö Alarm detected on module, relay triggered
Measure
Ö Relay triggered
Maintenance
Ö In Maintenance mode, relay is triggered
x
The "Type" fields are used to control (NC) or not (NO) the relays when alarms are OFF.
x
"Test" fields are used to manually control these relays.
12
JUNE 2009
Duplication prohibited
1.4.4
ESTEL Board
Environnement S.A
ESTEL CARD(S) Ö Remote controls
This screen displays the assignment of remote control inlets.
The available assignment choice are : « Inactive», « Stop mode », « Zero Ref. », « Zero », « Span »,
« Auto span».
«Test» column is used to display the value read at remote control inlet, for the selected assignment.
JUNE 2009
13
ESTEL Board
Environnement S.A
1.5
1.5.1
1.5.3
Duplication prohibited
INSTALLATION AND REPLACEMENT OF ESTEL BOARD
x
Switch off the analyzer and unplug the mains cable before any maintenance work of the
analyzer,
x
Respect connection of ESTEL board / MODULE board at J20 when reassembling.
Switch off the analyzer
1.5.2
Unplug the mains cable
Put off the cover
(2) Unscrew the screws located on lateral sides
(1) Unscrew the screws located on the rear panel
of the analyzer
(3) Lift up the cover
(4) Remove the cover by pulling it backward
14
JUNE 2009
Duplication prohibited
ESTEL Board
Environnement S.A
If the analyzer is already equipped with an ESTEL board, follow step 1.5.4.
If the analyzer is not equipped with ESTEL board, follow step 1.5.5.
1.5.4
Dismount ESTEL board
(1)
Module board
(2)
ESTEL board
(3)
J20 connector on Module board
(4)
(5)
Connecting cable between Estel board
/ Module board
Fixing screw of Estel board on rear
panel of the analyzer
Disconnect the connecting cable between ESTEL board (4) / Module board (3).
Unscrew the fixing screws (5) of ESTEL board on rear panel of the analyzer.
Remove ESTEL board.
Configure jumpers of the new board making functionality correspondences according to Table 1 or Table 2.
Re-assemble the board.
1.5.5
Unrivet the back plate (6) of rear panel of analyzer
Then, install at the same place, the
new plate (7) delivered with the board
(7)
JUNE 2009
15
ESTEL Board
Environnement S.A
1.5.6
Duplication prohibited
Installation of board inside the analyzer
ANALYZER
SWITCHED OFF
(1) Vertically insert the board inside its slot.
(2) Re-screw the board on the slot
(3) Fit again connector on ESTEL board
(4) Then, re-connect on Module board at J20
(5) Replace cover on the analyzer. See 1.5.3 .
(6) Connect mains cable and switch on the analyzer. See 1.5.2 and 1.5.1.
16
JUNE 2009
Duplication prohibited
1.6
ESTEL Board
Environnement S.A
OPTION OF EXTERNAL CONNECTION
Five different options of ESTEL external connection are available:
DESIGNATION
REFERENCE
Option of external Estel connection
P10-1337-A
MARK
Figure 3
x Cable
x D02-INF-37-37M-M-A
(1)
x Tie-point block interface board
x C10-0012-A
(2)
x DIN track
x G13-IB-18066
(3)
DESIGNATION
Option of external Estel connection + 4 insulated outputs.
REFERENCE
P10-1338-A
MARK
Figure 4
x Cable
x D02-INF-37-37M-M-A
(1)
x Tie-point block interface board
x C10-0012-A
(2)
x Symmetrical DIN track
Limit stop
x G13-IB-18066
D03-103-002-26
(3)
x 2-way galvanic insulator
x I11-Jk2000-2
(4)
DESIGNATION
REFERENCE
Option of external Estel connection + 1 insulated output
P10-1350-A
MARK
Figure 4
x Cable
x D02-INF-37-37M-M-A
(1)
x Tie-point block interface board
x C10-0012-A
(2)
x Symmetrical DIN track
Limit stop
x G13-IB-18066
D03-103-002-26
(3)
x 1-way galvanic insulator
x I11-Jk2000-1
(4)
DESIGNATION
Option of external Estel connection + 2 insulated outputs
REFERENCE
P10-1351-A
MARK
Figure 4
x Cable
x D02-INF-37-37M-M-A
(1)
x Tie-point block interface board
x
(2)
10-0012-A
x Symmetrical DIN track
Limit stop
x G13-IB-18066
D03-103-002-26
(3)
x 2-way galvanic insulator
x I11-Jk2000-2
(4)
DESIGNATION
Option of external Estel connection + 3 insulated outputs
REFERENCE
P10-1352-A
MARK
Figure 4
x Cable
x D02-INF-37-37M-M-A
(1)
x Tie-point block interface board
x C10-0012-A
(2)
x Symmetrical DIN track
Limit stop
x G13-IB-18066
D03-103-002-26
(3)
x 2-way galvanic insulator
1-way galvanic insulator
x I11-Jk2000-2
I11-JK2000-1
(4)
JUNE 2009
17
ESTEL Board
Environnement S.A
Duplication prohibited
Figure 3 – Option of external connection P10-1337-A
Figure 4 – Option of external connection + 4 insulated outputs P10-1338-A
18
JUNE 2009
SOREL Board
BOARD OF LOGIC INPUTS / OUTPUTS
OPTION OF 2M ANALYZERS
- April 2010 -
WARNING
Information contained in this document are likely to be modified without notice.
The designer reserves the right to modify the equipment without improving this document,
therefore, information of this document does not represent a commitment under ENVIRONNEMENT S.A.
ENVIRONNEMENT S.A. all right reserved.
111 bd Robespierre, 78300 POISSY - -TEL. 33(0)-1.39.22.38.00 – FAX 33(0)-1.39 65.38.08
http://www.environnement-sa.com
SOREL Board
Environnement S.A
Duplication prohibited
SOREL BOARD
1.1
FUNCTION AND USE
3
1.2
TECHNICAL CHARACTERISTICS
3
1.3
CONFIGURATION
4
1.3.1
Programmation
5
1.3.2
CONFIGURATION Ö Relays and remote controls
6
1.3.3
TESTS Ö ESTEL card
7
1.4
INSTALLATION OR REPLACEMENT OF SOREL BOARD
8
1.4.1
Switch off the analyzer
8
1.4.2
Unplug the mains cable
8
1.4.3
Put off the cover
8
1.4.4
Dismount SOREL board
9
1.4.5
Unrivet the back plate (6) of rear panel of analyzer
9
1.4.6
Installation of board inside the analyzer
10
Table 1 - Configuration of SOREL board
4
Figure 1 – SOREL board
4
Up-to-date:
Pages
Up-to-date :
1
10-04
2
10-04
3
05-04
4
10-04
5
05-04
6
05-04
7
05-04
8
05-04
9
05-04
10
05-04
2
APRIL 2010
SOREL Board
Duplication prohibited
1.
Environnement S.A
SOREL BOARD
SOREL board is a universal board of logic inputs/outputs for the 2M analyzers. It is optional. It is
possible to install up to 2 SOREL boards in an analyzer.
1.1
FUNCTION AND USE
SOREL board has 2 functions:
x
Relays control ( 4 in all )
x
Remote controls ( 4 inputs )
SOREL board communicates with measurement module through Bus i2C and relieves it of the
Inputs/Outputs functions. It enables the remote control and/or the remote signaling of certain functions
as: "measurement", "zero", "calibration" and "alarm".
1.2
TECHNICAL CHARACTERISTICS
Management by specialized micro controller:
x
only one power supply of 24 volts,
x
4 logic inputs,
x
4 contacts for remote signaling, potential configurable by the user,
x
Visualization of i2C communication using a LED.
Electric connection:
x
4-point connector for link with module boards of 2M series,
x
Inputs / Outputs centralized on plug-in connector. This connector in screwed on the rear
panel of the analyzer using a back-plate.
Voltage and current on relays:
x
Maximum voltage by relay contact: 50 volts D.C.
x
Maximum current by relay contact : 1 Ampere at 24 V D.C. (resistive load)
Voltage at logic inputs:
x
MAY 2004
Maximum voltage 24 V D.C.
3
SOREL Board
Environnement S.A
1.3
Duplication prohibited
CONFIGURATION
Table 1 - Configuration of SOREL board
JUMPERS MARK
SYMBOLS
NATURE OF OPERATIONS
SOREL selection, board N° 1
SOREL selection, board N° 2
SOREL selection, board N° 3
SOREL selection, board N° 4
ST1, ST2, ST3
SOREL selection, board N° 5
SOREL selection, board N° 6
SOREL selection, board N° 7
SOREL selection, board N° 8
0 V to ground (default)
ST4
Floating 0 V
SW1 Relay nb 1
SW2 Relay nb 2
SW3 Relay nb 3
SW4 Relay nb 4
Potential free contact
Referenced contact to 0 V and 24V
18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
3 Remote control 1
1 +24 V
4 GND
(2) 5 to 24V D.C.
2
(1)
Relay contact #4 (-)
Relay contact #4 (+)
Relay contact #3 (-)
Relay contact #3 (+)
Relay contact #2 (-)
Relay contact #2 (+)
Relay contact #1 (-)
Relay contact #1 (+)
GND
Remote control #4
GND
Remote control #3
GND
Remote control #2
GND
Remote control #1
+15V
+24V
3 Remote control 1
+
Figure 1 – SOREL board
NOTE:
Output relay contacts are normally open when analyzer is switched off.
4
APRIL 2010
SOREL Board
Duplication prohibited
1.3.1
Environnement S.A
Programmation
The hereafter screens (§ 1.3.1 à § 1.3.3) are given as example.
Refer to technical manual of the analyzer in which SOREL board is installed.
The analyzer automatically detects the presence of one or several SOREL and/or ESTEL boards and
offers menus enabling the user to adjust and configure each board.
x
In CONFIGURATION menu of main software program, the "Analog outputs", "Analog
inputs", "Relays and remote controls" items are only displayed if the SOREL and/or ESTEL
board option is available. Only, the sub-menu "Relay and remote controls is necessary
to program the SOREL board.
x
In the TESTS menu of the main program, the item "ESTEL card" is displayed if one
SOREL board at least is detected.
The same screen as for ESTEL board must be used, but it is necessary not to take into
account of data about analog inputs and outputs.
MAY 2004
5
SOREL Board
Environnement S.A
1.3.2
Duplication prohibited
CONFIGURATION Ö Relays and remote controls
This screen enables to configure function of each input / output of SOREL and/or ESTEL board(s).
x
SOREL board is displayed as an ESTEL board,
x
The "ESTEL card Nb: " is used to select what board to configure.
x
"Relays" fields are used to control the relays according to each analyzer: refer to
CONFIGURATION Ö Relays and remote controls paragraph of the technical manual of your
analyzer.
x
The "Type" fields are used to program relays into "normally closed" (NC) or "normally open"
(NO) when alarms are OFF.
x
The "Mode" field is used to configure the working mode of remote controls.
Two different modes are possible:
"State" mode: control is activated as long as remote control is active (closed contact).
"Rise" mode: control is activated when state modification of remote control is detected. When it is
down, control remains active. A new modification of state de-activates control.
6
MAY 2004
SOREL Board
Duplication prohibited
1.3.3
Environnement S.A
TESTS Ö ESTEL card
This screen is used to check operation of remote controls and relays.
Analog functionalities are not active for SOREL board.
The “Estel card Nb:” field is used to select the board to be tested.
The “Out” fields are used to control the relays manually.
The "Rem." fields are used to know state of these logic inputs.
Definition of the specific keys of this screen:
MAY 2004
0/OFF
Opens all the relay contacts.
4000/ON
Closes all the relay contacts.
7
SOREL Board
Environnement S.A
1.4
1.4.1
1.4.3
Duplication prohibited
INSTALLATION OR REPLACEMENT OF SOREL BOARD
x
Switch off the analyzer and unplug the mains cable before any maintenance work in the
analyzer,
x
Respect connection of SOREL board / MODULE board at J20 when reassembling.
Switch off the analyzer
1.4.2
Unplug the mains cable
Put off the cover
(2) Unscrew the screws located on lateral sides
(1) Unscrew the screws located on the rear panel
of the analyzer
(3) Lift up the cover
(4) Remove the cover by pulling it backward
8
MAY 2004
Duplication prohibited
SOREL Board
Environnement S.A
If the analyzer is already equipped with a SOREL board, follow step 1.4.4
If the analyzer is not equipped with ESTEL board, follow step 1.4.5
1.4.4
Dismount SOREL board
(1)
Module board
(2)
SOREL board
(3)
Connector J20 on Module board
(4)
(5)
Connecting cable between Sorel board
/ Module board
Fixing screw of Sorel board on rear
panel of the analyzer
Disconnect the connecting cable between SOREL board (4) / Module (3) board.
Unscrew the fixing screws (5) of SOREL board on rear panel of the analyzer.
Remove SOREL board.
Configure jumpers of the new board carrying out correspondence of functionalities according to Table 1.
Re-assemble the board.
1.4.5
Unrivet the back plate (6) of rear panel of analyzer
Then, install at the same place the
new plate (7) delivered with the board
(7)
MAY 2004
9
SOREL Board
Environnement S.A
1.4.6
Duplication prohibited
Installation of board inside the analyzer
ANALYZER
SWITCHED OFF
(2) Re-screw the board on back plate
(1) Vertically insert the board inside its slot.
(3) Fit again connector on SOREL board
(4) Then, re-connect on Module board at J20
(5) Replace cover on the analyzer. See 1.4.3.
(6) Connect mains cable and switch on the analyzer. See 1.4.2 and 1.4.1.
10
MAY 2004

advertisement

Related manuals

Download PDF

advertisement