CAIRE Hi Flow Stroller Operating Instructions
Реклама
Реклама
PATIENT
OPERATING
INSTRUCTIONS
Hi Flow
Hi Flow Stroller
Warning Information
Read this page carefully before using your unit.
• The unit contains liquid oxygen which is extremely cold; almost 300 degrees below zero Fahrenheit.
Exposure to such a low temperature can cause severe frostbite.
• Liquid and gaseous oxygen, though nonflammable, cause other materials to burn faster than normal.
This hazard, along with the low temperature of liquid oxygen, warrants certain safety precautions.
• Do not use or store your unit in an area where combustible materials such as oils, greases, aerosol sprays, lotions or solvents are present.
•
Do not smoke while operating unit.
• Do not use or store your unit within five feet of electrical appliances, especially stoves, heaters, toasters, and hair dryers.
• Keep your unit in a well-ventilated area.
• In the event of an accidental tip-over, immediately but cautiously return unit into an upright position.
If any liquid oxygen is escaping, leave the area immediately and call your health care provider.
•
Do not touch the frosted parts of any unit.
• Do not store or operate the portable coupled to the
Liberator/Low Loss.
• Applicable EMC information can be found in the Hi
Flow Stroller Service Manual at www.cairemedical.com.
Explanation of Other Symbols
Name and address of manufacturer
Authorized representative in the European Community
Caution, consult accompanying documents
Non-flammable gas
Oxidizing substances
Type B Applied Part (degree of protection against electric shock)
Patient’s Name
Prescribed Oxygen Flow Setting
Doctor’s Telephone Number
Distributor’s Name
Person to Contact
Distributor’s Emergency Telephone Number
Special Instructions
Patient’s Records
1
Hi Flow Stroller
Table Of Contents
Introduction _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ 3
Controls _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _4–5
Filling Instructions _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ 6–9
Operating Instructions _ _ _ _ _ _ _ _ _ _ _ _ _ 10–14
Additional Information _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ 15
Patient Notes _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ 16
Specifications:
• Mode of Operation: Continuous
• Type of Protection Against Electric Shock: Internally
Powered Equipment
• Degree of Protection Against Electric Shock: Type B
Applied Part
• Classification according to the degree of protection against ingress of water: IPX0 - Ordinary Equipment
• Equipment is not suitable for use in the presence of flammable mixtures
2
Important: Read this manual thoroughly before
operating the Hi Flow Stroller.
Rx only.
Warning: This device is not intended for life
sustaining use.
Caution: Use the Hi Flow Stroller only as directed by
your doctor.
Warning: If you feel the equipment is not working
properly, call your health care provider. Do not attempt to repair or adjust the units yourself.
Introduction
The liquid oxygen system includes a stationary unit and the Hi Flow Stroller, which provides supplementary oxygen as prescribed by your physician. This POI contains the instructions for using the Hi Flow Stroller.
Refer to the POI supplied with the stationary unit for its operation.
The Liberator is the larger container and is intended for stationary use. You may take oxygen directly from the
Liberator. The unit is filled by your health care provider.
The Hi Flow Stroller provides an ambulatory source of oxygen for an extended period of time. They are filled from the Liberator.
Hi Flow Stroller
3
Hi Flow Stroller
Controls
1. Electronic Liquid Level Gauge
2. Flow Control Knob
3. Oxygen Tube Connector
4. Vent Valve
5. Fill Connections (see next page for QDV
Identification)
6. Carrying Strap with Shoulder Pad
7. Condensation Pad
2
6
1
3
5
4
7
4
Quick Disconnect Valve (QDV) Identification
Hi Flow Stroller
Portable
Connector:
Stationary
Connector:
QDV Style: CAIRE
Side Fill
SF Locking
LINCARE
(UC)
CRYO2 LIFE-OX
TF Locking
PENOX TAEMA
TF Push-on
PB
5
Hi Flow Stroller
To Fill the Top Fill Hi Flow Stroller
1. Clean the fill connectors on both the stationary and the Hi Flow Stroller with a clean, dry, lint-free cloth.
Caution: The fill connectors must be dry, because
moisture can cause your Hi Flow Stroller and stationary unit to freeze together.
2. Make sure the flow control is in the off
(0) position.
3. Position the fill connector in the upright position over the stationary unit’s connector.
4. a. If your connector is a push-on:
Step 3
Lower the Hi Flow Stroller unit on to the stationary until you feel the connector engage.
b. If your connector is Cryo2:
Rotate Hi Flow Stroller until you feel the connectors engage. Then, carefully and firmly, rotate the Hi Flow Stroller counterclockwise ( ) while pressing downward.
c. If the connector is a Penox, Life-Ox, or Lincare:
Rotate Hi Flow Stroller until you feel the connector engage. Then, carefully and firmly, rotate the Hi
Flow Stroller clockwise ( ) while pressing downward.
Note: Continued on step 6 (page 9).
6
Step 4a Step 4b Step 4c
To Fill the Side Fill Hi Flow Stroller
1. Clean the fill connectors on both the stationary and the Hi Flow Stroller with a clean, dry, lint-free cloth.
Caution: The fill connectors must be dry, because
moisture can cause your Hi Flow Stroller and stationary unit to freeze together.
2. Make sure the flow control is in the off (0) position
3. Make the Hi Flow Stroller fill connector in the upright position to the stationary connector.
4. Rotate the Hi Flow Stroller counterclockwise ( ) until you feel the pin and slot engage.
5. Carefully and firmly rotate the Hi Flow Stroller back to the upright position. Now the units are locked together.
Note: Continued on step 6.
Hi Flow Stroller
Step 3
Step 4 Step 5
7
Hi Flow Stroller
Note: For push-on style connector, hold the unit
in place throughout the fill.
6. Open the vent valve to begin filling your Hi Flow
Stroller. You should hear some oxygen escaping; do not worry. The hissing sound you hear is perfectly normal.
Note: Liquid level gauge readings are most
accurate 30 seconds or more after the Hi Flow
Stroller is filled.
7. You may also see vapor around the connection. This is normal and is due to the great difference in temperature between the liquid oxygen and the warm air surrounding the units.
8. When the unit is full, close the vent valve. You can tell that the unit is full when the hissing sound changes in tone, and some liquid comes out of the vent line.
Note: Closing the valve after a nearly continuous
stream of liquid is discharging fills the Hi Flow
Stroller most completely, but wastes some liquid oxygen.
Step 6
8
Closing the valve at the first sign of liquid discharge does not fill the Hi Flow Stroller as completely, but saves liquid oxygen.
9. For the push-on style connectors, push the release button down until the units separate. For all other connectors rotate the unit until it separates from the stationary unit.
Caution: Check liquid level gauge only after the vent
valve is closed.
Caution: If the Hi Flow Stroller does not separate
easily, do not use force. The units may be frozen together. Wait until the units warm up — then they’ll separate easily. Do not touch any frosted parts.
Should there be any liquid leakage from the stationary unit or Hi Flow Stroller after separation, reattach the Hi Flow Stroller immediately. Again separate the units. If the problem persists, leave units coupled and call your health care provider.
Hi Flow Stroller
Warning: Do not operate Hi Flow Stroller while
attached to the stationary unit.
Warning: Should the leakage be excessive to the
point that a stream of liquid is present, leave the area and call your health care provider immediately.
Caution: If the Hi Flow Stroller is being refilled
immediately after a period of use, fill unit, set FCV to zero, and then wait 10 minutes for pressure to stabilize before use.
Warning: Liquid or cold gaseous oxygen can frostbite
the skin. Oxygen will result in increased fire hazard.
Portable Release
Button (push-on style only)
9
Hi Flow Stroller
Operating Instructions
1. Push the cannula breathing tube firmly on to the oxygen tube connector.
2. Adjust your cannula breathing tube to the proper position so that you will be able to breathe the oxygen comfortably.
3. Turn the flow control knob on top of the Hi Flow
Stroller clockwise until the prescribed flow rate
(numeral) is visible in the knob “window” and a positive detent is felt.
Warning: Out-of-specification oxygen flow will result
if the flow control knob is set between flow rates.
4. You should be receiving oxygen now. To stop flow of oxygen, turn the flow control knob counterclockwise to the off (0) position.
10
Caution: Always turn the
flow control knob off (0) when not in use.
*Cannula is not included
Hi Flow Stroller
5. Use the following chart as a guideline to determine the length of time the Hi Flow Stroller will operate:
Model Off 0.50 1.0
Hi Flow Stroller Nominal 71:00 32:00 16:00
2.0
8:00
FCV Setting
2.5 3.0
6:20 5:20
4.0 6.0
4:00 2:40
8.0 10.0 12.0
2:00 1:40 1:20
15.0
1:00
Note: Times are in hours and minutes (format 00:00).
Note: The “Nominal” times are for ideal conditions, i.e. maximum fill, exact flow rates, good loss rate, unit not being moved, etc.
These times are the maximum expected.
Note: Your individual results will vary.
11
Hi Flow Stroller
6. To verify the level of liquid oxygen in the unit with the electronic liquid level gauge:
• Depress the push button on top of the unit for two seconds minimum. Read the arc of LEDs which indicate contents level.
Caution: The Hi Flow Stroller is empty if only the first
LED on the far left is lit.
• If the Low Battery indicator lights up when the button is depressed, notify your health care provider.
Note: The Hi Flow Stroller will continue to supply
oxygen even if the Low Battery Indicator lights up, as long as there is liquid oxygen in the unit.
12
7. Under certain environmental conditions, with continuous use, and especially with the portable not being moved, it may develop excessive frost around the vaporizer coil and on the case. You may reduce this frost by tapping the unit and/or wiping any accumulated frost off the case.
WARNING: Due to the possibility of excessive
moisture / ice build up it is recommended that the Hi
Flow Stroller remain idle for 1 hour after use to decrease moisture buildup.
Hi Flow Stroller
13
Due to the higher flow rates associated with the Hi Flow Stroller, condensation and some frosting will occur on the unit, particularly on flow settings of 10 LPM and higher. It is recommended that the unit be allowed at least 1 hour of inactivity after use to decrease the possibility of excessive build up due to continuous usage.
Hi Flow Stroller
8. After the unit is empty and has warmed up to room temperature, remove any moisture from the condensation cup:
• Unzip soft feature case access door at bottom of unit.
• Remove the pad and wring out any absorbed moisture.
Allow the pad to dry completely before replacing it back in the unit.
Note: If heavily soiled, the pad can be machine
washed (with bleach to disinfect).
• Place dry pad back in the unit. Make sure to completely re-zip the zipper.
14
Cleaning Standard
Warning: Clean only after unit is empty and vented.
• Clean by applying household glass cleaner to a lintfree cloth and wipe external surfaces – do not get glass cleaner on any internal components or valves.
• Allow unit to dry thoroughly before using.
Note: Note to health care provider – for reprocessing
procedure, see applicable service manual.
WEEE and RoHS
Hi Flow Stroller
Our products will comply with the Restriction of
Hazardous Substances (RoHS) directive. They will not contain more than trace amounts of lead or other hazardous material content.
Accessories
The lightweight Stroller carrying cart weighs only
3-1/4 lbs. The cart accommodates both the
Stroller or the Sprint.
This symbol is to remind the equipment owners to return it to a recycling facility at the end of its life, per Waste Electrical and
Electronic Equipment (WEEE) Directive.
15
Patient Notes
16
2 2 0 0 A i r p o r t I n d u s t r i a l D r i v e
S t e 5 0 0
B a l l G r o u n d , G A 3 0 1 0 7
U . S . A .
w w w . c a i r e m e d i c a l . c o m
M . D . D . R e p re s e n t a t i v e : M e d i c a l P ro d u c t S e r v i c e s