Mettler Toledo transmitter M 700(X) - additional functions Operating Instruction
Add to My manuals56 Pages
Mettler Toledo transmitter M 700(X) - additional functions offers a wide range of functionalities to enhance your process analysis capabilities. With its modular design, you can customize your device to meet your specific measurement needs. The additional functions provide advanced features such as data logging, tolerance band monitoring, and concentration determination, enabling you to optimize your processes and ensure accurate and reliable measurements. Explore the various options available to unlock the full potential of your M 700(X) transmitter.
advertisement
![Mettler Toledo transmitter M 700(X) - additional functions Operating Instruction | Manualzz Mettler Toledo transmitter M 700(X) - additional functions Operating Instruction | Manualzz](http://s1.manualzz.com/store/data/039327227_1-85a5e204dbe34f365f7ec0078a0c53a7-360x466.png)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 1 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW
M 700(X)
Additional functions
O pt ion retrofitt ab le
Latest product information: www.mtpro.com
Order number: 52 121 241
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 2 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW
Warranty
Defects occurring within 1 year from delivery date shall be remedied free of charge at our plant
(carriage and insurance paid by sender).
©2005 Subject to change without notice
Return of products under warranty
Please contact your local Mettler-Toledo representative before returning a defective device. Ship the cleaned device to the address you have been given. If the device has been in contact with process fluids, it must be decontaminated/disinfected before shipment. In that case, please attach a corresponding certificate, for the health and safety of our service personnel.
Disposal
Please observe the applicable local or national regulations concerning the disposal of
“waste electrical and electronic equipment”.
Registered trademarks
The following registered trademarks are used in this instruction manual without further marking
CalCheck
Calimatic
Sensocheck
Sensoface
ServiceScope
VariPower
SMARTMEDIA ® is a registered trademark of Toshiba Corp., Japan
InPro ® is a registered trademark of Mettler Toledo GmbH, Switzerland
Mettler-Toledo GmbH, Process Analytics, Industrie Nord,
CH-8902 Urdorf, Tel. +41 (44) 736 22 11 Fax +41 (44) 736 26 36
Subject to technical changes. Mettler-Toledo GmbH, 11/05.
Printed in Germany.
TE-201.011-MTE06 071105
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 3 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW
Additional functions
for the M 700(X) modular measuring system
Additional functions
Latest product information ...........................................................................4
How to order an additional function ............................................................5
Information on TAN release and SmartMedia cards
Activating an additional function (Option release) ........................................6
SmartMedia card ........................................................................................30
Provided by: TAN
52121198 SW 700-001 KI recorder .........................................................7
52121199 SW 700-002 Buffer sets to be entered (pH) ..........................13
52121201 SW 700-004 ServiceScope (pH).............................................14
52121202 SW 700-005 Tolerance band recorder (pH)............................15
52121203 SW 700-006 Current characteristic definable ........................17
52121204 SW 700-008 TC ultrapure water (Cond) ...............................19
52121205 SW 700-009 Concentration determination (Cond) ................20
52121250 SW 700-011 High CO
2 compensation (O
2
) ...........................26
52121274 SW 700-012 ISFET for ISM module pH 2700i(X)....................28
Provided by: TAN. SmartMedia card required.
52121192 SW 700-102 5 loadable parameter sets ................................35
52121193 SW 700-103 Measurement recorder .....................................37
52121194 SW 700-104 Extended logbook ............................................40
52121195 SW 700-106 Software update ..............................................41
52121196 SW 700-107 AuditTrail and Signature ...................................45
3
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 4 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW
Latest product information
Additional functions
The M 700 is an expandable modular process analysis system.
For latest product information, please refer to: www.mtpro.com
Additional functions
A current version of the respective function description is available as pdf for download.
4
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 5 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW
How to order an additional function
Device-specific TAN (transaction number)
Additional functions expand the device capabilities.
The additional functions are device-specific. When ordering an additional function, you therefore have to specify the serial number and hardware version of your M 700 FRONT in addition to the respective order number.
The manufacturer then supplies a TAN (transaction number) to release the additional function.
Serial number of M 700 FRONT
Menu Display Serial number FRONT M 700
Menu selection
0.002 mS/cm
23.7 °C
Menu selection
Call up diagnostics.
From the measuring mode:
Press menu key to select menu.
Select diagnostics using arrow keys, confirm with enter.
Select:
Return to meas
[enter]
Lingua
0.003 mS/cm
23.7 °C
Diagnostics
Message list
Logbook
Device description
FRONT M 700-011
BASE M 700-021
Module Cond Ind 7700
Return Set favorite
0.003 mS/cm
22.7 °C
Device description
FRONT M 700-011
Operating panel M 700
Hardware: 2, Software: 6 .0
Serial number: 0000815
Diagnostics
Select Device description using arrow keys, confirm with enter.
Device description
Please specify the serial number/hardware version when ordering an additional function.
Module FRONT BASE
Return
5
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 6 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW
Activate additional function
Select menu: Parameter setting/System control/Release of options
Note: The TAN for releasing an additional function is only valid for the device with the corresponding serial number (see previous page).
Menu Display Activate additional function
6
Menu selection
7.00 pH
25.6 °C
Select:
Return to meas
[enter]
Lingua
25.6 °C
0.003 mS/cm
Parameter setting
Viewing level
Operator level
Admin. level
(All Data) view
(Operation Data) opl
(All Data) adm
Menu selection
Call up parameter setting.
From the measuring mode:
Press menu key to select menu.
Select parameter setting using arrow keys, confirm with enter.
Parameter setting
Select Administrator level using arrow keys, confirm with enter.
Enter passcode and confirm
(Passcode as delivered: 1989).
Return
16.9 %Air
25.6 °C
Parameter setting (Administrator)
System control
FRONT M 700-011
BASE M 700-021
Module O2 4700
Module pH 2700
Module Cond 7700
Return
0.003 mS/cm
25.6 °C
Release of options (Administrator)
SW 700-001 KI recorder
SW 700-002 Buffer sets
SW 700-003 Adaptive timer
SW 700-004 ServiceScope
SW 700-005 Cal tolerance band
SW 700-006 Current characterist.
Return
Active
Inactive
Inactive
Inactive
Inactive
Inactive
Select System control using arrow keys, confirm with enter.
Then select Release of options using arrow keys, confirm with
enter.
Release of options
Select the additional function to be released.
Set option to “active”. Enter the
TAN at the prompt. (Note: The TAN only applies for the device with the corresponding serial number, see previous page.) The option is available after the TAN has been entered.
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 7 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW
SW 700-001: KI recorder
Select menu: Parameter setting/FRONT M 700/KI recorder
Additional function SW 700-101
The KI recorder follows the course of the process and releases a messages in the event of a deviation. It is called up from the measurement recorder: meas.
Continuous processes
The measured signal is surrounded by a dynamically tracked, processcontrolled tolerance band. Slight fluctuations are tolerated. If the measured value leaves the tolerance range, a message is released (Fig.).
The recording interval can be set from 10 s to 60 h.
Variable Message
Tolerance band around the measured signal generated by a low-pass filter
Measured signal
Fig.: KI recorder, continuous process t
Batch processes
The KI recorder provides an automatic self-learning function. This allows checking whether regularly recurring processes keep the preset tolerances.
Deviations from the normal course of process (plus preset tolerances) generate a message (Fig.). Maximum process duration can be set from 1 to 60 h.
Variable Start End
Preset tolerance band around the measured signal
Measured signal
Fig.: KI recorder, Batch process (Self-learning function) t
7
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 8 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW
SW 700-001: KI recorder
Note: Function check is active during parameter setting.
Menu Display Configure KI recorder
5 pH 9 pH
22 °C 32 °C pH
8,56 pH, 29.4
°C
09.07.02 11:23:33
KI recorder
Info on KI recorder
Recording data
Wait - Function check
Failure right value
Failure left value
Keys :
[enter]
[meas]
[menu]
Cursor to current
To meas. display
Call up menu
Info
°C
Cursor 1 line
Cursor 1/2 page
Close
KI recorder
Called up directly from the measuring mode:
Press meas key.
For function description, press
“Info” softkey
“Info” softkey
Explains icons and capabilities of the
KI recorder
7.00 pH
25.6°C
KI recorder (Administrator)
KI recorder
Channel 1
Channel 2
Off
Continuous process
Batch process
Recording interval
Monitoring
Return
10s
Failure
Configure KI recorder
• Parameter setting FRONT M 700
• Select KI recorder
• Select the variables to be displayed as well as the start and end values
• Set recording interval
(10 s to 60 h) (per pixel)
• Monitoring
(Off/Failure/Maintenance request)
7.00 pH
25.6 °C
Function control matrix (Administrator)
ParSet KI rec.
Fav
– Input OK2
Left softkey
Right softkey
Profibus DO 2 –
–
–
–
–
Return Connect
Control KI recorder
Example: Configure softkey for
KI recorder start/stop:
• Select parameter setting, then:
• System control
• Function control matrix
• Softkey: “KI rec. Start/Stop”
8
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 9 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW
SW 700-001: KI recorder
Menu Display KI recorder: Batch process
7.00 pH
25.6°C
KI recorder (Administrator)
KI recorder
Monitoring
Off
Channel 1 Continuous process
Channel 2
Max. process duration
Batch process
New record Yes No
Failure
Return
Configure KI recorder
• Parameter setting: “Module FRONT”
• Select KI recorder
• Select the variables to be displayed as well as the start, end values, and deviation (%)
• Set process duration
(200 samples are recorded)
• Select “New record Yes”
• Monitoring
(Off/Failure/Maintenance request)
Call up KI recorder
From the measuring mode: Press meas.
Parameters adjustable for batch operation
The gray envelope represents the deviation [%]
End
Variable
Start
9
I pH
°C
I
5
25
15
50 100 150 200
Channel 1
Channel 2
0
Record
50 100 150
KI stop
200
Max. process duration
When the parameters are edited once more, the reference curve is automatically re-adjusted to start, end, and deviaton.
When variable or process duration are changed, the curve will be deleted!
9
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 10 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-001: KI recorder
Display
9
I pH
5
27
°C
I
20
50
0
KI pause
50
100 150
100 150
KI stop
200
200
KI recorder: Batch process
Record reference curve
Start “Record” with left softkey. Reference value and envelope are displayed during the recording process. At the end of recording or after “KI stop” the KI recorder switches to control mode.
9
I pH
5
27
°C
I
50
20
0
KI start
50
100 150 200
In control mode only the envelope is visible.
9
I pH
5
27
°C
I
50
20
0
KI pause
50
100
Cursor
150
KI stop
200
100 150
100 150
KI stop
200
200
Batch control
After a synchronous start the recorder monitors whether the currently measured values remain within the envelope. Every deviation can trigger a message.
KI pause: Stops the recorder.
KI stop: Terminates control.
Pressing “Continue” restarts a stopped recorder.
Note: During “function check” (e.g. during calibration) neither recording nor checking of a batch are interrupted. The last measured value continues to be used.
10
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 11 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-001: KI recorder
Display KI recorder: Batch process
9
I pH
5
50
Evaluation channel 1:
End 9.00 pH
Start
Deviation
Process duration
5.00 pH
10 %
10 h
KI pause
100 150 200
Max
Min
Max. deviation
Messages
KI stop
8.35 pH
5.80 pH
7 %
0x
°C
I
9
I pH
5
27
20
50 100
0 50
Control via OK2
100
150
150
200
200
7.00 pH
25.6 °C
Contact K3
Usage
Contact type
NAMUR maintenance!
NAMUR function check
Limit
Rinse contact
KI rec active
Return
Evaluation
Within the KI recorder function you can evaluate a reference curve directly on the device.
To do so, use the arrow keys (Up/Down). The reference curves and the most important parameters are displayed one after the other.
External control
The Start/Stop function can be remote-controlled via the OK2 input (BASE module) or via the PA 700 (X) PROFIBUS module.
(To set the control element, call up:
Parameter setting / System control / Function control matrix).
The softkeys will then be deactivated.
The K1, K2, and K3 contacts of the BASE module can be assigned to the “KI rec active” signal.
• Save / load reference curves
Up to 5 batch reference curves can be stored on and retrieved from a SmartMedia card:
Parameter setting / System control / KI batch recording.
The reference curve, envelope, time response, and KI settings are stored.
• Archiving of completed batch processes
Complete batch results can be archived on the SmartMedia card. All reference values and current batch data are stored in the archive.
11
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 12 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
12
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 13 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-002: Buffer set to be entered (pH)
Select menu: Parameter setting/System control/Buffer table
Individual buffer set (with 3 buffer solutions) for pH measurement
Buffer table
You can enter an individual buffer set (with 3 buffer solutions).
To do so, you enter the nominal buffer values for the correct temperature
(range 0 ... 95 °C, 5°C steps). Then this buffer set is available in addition to the permanently set standard buffer solutions in the “Calimatic buffer” menu (select “Table”).
Menu Display
7.20 pH
25.6 °C
System control (Administrator)
Calculation Blocks
Point of measurement
Release of options
Logbook
Buffer table
Concentration chart
Return
Buffer table (Administrator)
Buffer 1
Buffer 2
Buffer 3
7.10 pH
25.6 °C
Buffer table: Enter values
Enter buffer set
• Call up parameter setting
• System control
• Select “Buffer table”
• Select buffer to be entered
Return
7.10 pH
25.6 °C
Buffer 1 (Administrator)
Nominal buffer value pH value at 00 °C pH value at 05 °C pH value at 10 °C pH value at 15 °C pH value at 20 °C
Return
+04.00 pH
+04.00 pH
+04.00 pH
+04.00 pH
+04.00 pH
+04.00 pH
• Enter nominal buffer value and all other values for the correct temperature (right/left arrow keys to select position, up/down arrow keys to edit number, confirm with
enter.)
The special buffer set is selected as follows:
Parameter setting/Module pH/Cal preset values/Calimatic buffer/Table.
13
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 14 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-004: ServiceScope (pH)
Select menu: Diagnostics/Module pH 2700/ServiceScope
ServiceScope
ServiceScope monitors the pH input signal and checks whether it remains in the control range. An error message is generated if the input signal exceeds the failure limit.
Menu Display
ServiceScope
Fail
Noise level
Return
ServiceScope
7.00 pH
0.2 °C t
ServiceScope
(Diagnostics/Module pH 2700)
Displays the noise levels over the time. Allows distinction of individual disturbances, periodic and broadband disturbances, which is helpful for troubleshooting. An error message is generated if the noise level exceeds the failure limit.
14
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 15 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-005: Tolerance band recorder (pH)
Select menu: Diagnostics/Module pH/Tolerance band recorder
Tolerance band recorder
(Adjustment of tolerance band: see following page)
Tolerance band calibration prevents that slight calibration scatter of zero and slope immediately leads to a readjustment of the calibration data and thus to a shifting of the measured value. If the zero and slope values determined during calibration remain within their tolerance bands, the new data are not stored. If one of the calibration values lies outside the tolerance band, the new data are taken over.
Drift due to aging or calibration scatter can be identified at a glance, thus allowing to draw conclusions as to electrode life and the required calibration interval.
In the Calibration record you see whether the data have been stored or whether the old data can still be tolerated.
Menu Display
Tolerance band recorder
Cal record
6.00
T-band recorder
ZERO 8.00
50.7
Return
7.00 pH
22.3 °C
Slope 61.7
Tolerance band recorder
(Diagnostics/Module pH 2700)
Records the tolerance ranges for zero and slope over the time. New calibration values are only accepted if the tolerance limits have been exceeded.
Display can be graphical or as a listing.
7.00 pH
22.3 °C
Cal record
Last calibration
Cal tolerance band
Cal mode
Zero point
Slope
Isothermal potential
Return
01.05. 2003 11:24
Old el data
7.00 pH
58.0 mV/pH
0 mV
Cal record
(Diagnostics/Module pH 2700)
In the Cal record you can see whether the data have been stored
(New el data) or whether the old calibration data can still be tolerated
(Old el data).
15
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 16 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-005: Tolerance band recorder (pH)
Adjustment of tolerance band. Select menu:
Parameter setting/Module pH 2700/Cal preset values/Cal tolerance band
Menu Display
Menu selection
Select:
Return to meas
7.0 pH
17.1 °C
[enter]
Lingua
Parameter setting
Menu selection
From the measuring mode:
Press menu key to select menu.
Select parameter setting using arrow keys, confirm with enter.
Passcode as delivered: 1989
7.00 pH
17.1°C
Parameter setting (Administrator)
System control
FRONT M 700-011
BASE M 700-021
Module Cond 7700
Module PID 700
Module pH 2700
Return
7.00 pH
17.1 °C
Module pH 2700 (Administrator)
Input filter
Sensor data
Cal preset values
TC process medium
ORP/rH value
Delta function
Return Block
7.00 pH
17.1 °C
Cal tolerance band (Administrator)
Calibration data taken over when tolerance band exceeded
Tolerance band check
Tolerance Zero
Tolerance Slope
On Off
+02.00 pH
002.0 mV/pH
Return Block
Parameter setting
Select “Module pH 2700”, confirm with enter.
pH module
Select “Cal preset values” using arrow keys.
Confirm with enter.
Then select “Cal tolerance band”.
Cal tolerance band
Enter tolerance ranges for zero and slope
(enter numbers using arrow keys, confirm with enter.)
16
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 17 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-006: Current characteristic definable
Select menu: Parameter setting/BASE M 700 (Out)/Output current/Curve
Menu Display
7.03 pH
25.6 °C
Output current I1 (Administrator)
Variable
Curve
Output
Start
End
Output filter
Abort pH
Linear
Trilinear
Function
Chart
0000 s
OK
7.03 pH
25.6 °C
Output current I1 (Administrator)
Variable
Curve pH
Chart
Chart
Output 4 ... 20 mA
Output filter 0000 s
Behavior during messages
Abort OK
Specify current characteristic
Chart characteristic
Assign output current to measured value in 1 mA steps.
Select menu: Parameter setting/
BASE M 700 (Out)/Output current
First select process variable!
Then curve: Chart
The “Chart” menu is displayed.
7.03 pH
25.6 °C
Chart (Administrator)
Value at 00 mA:
Value at 01 mA:
Value at 02 mA:
Value at 03 mA:
Value at 04 mA:
Value at 05 mA:
Abort
+ 00.00 pH
+ 00.00 pH
+ 00.00 pH
+ 00.00 pH
+ 00.00 pH
+ 00.00 pH
OK
The entered values must be continuously rising or falling.
On the following page, you find a listing to write down your adjustments.
17
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 18 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
Current curve values - Original for copy
Parameter setting/BASE M 700 (Out)/Output current/
Curve: Chart, individual settings
Value at measured variable
00 mA ............... _________
01 mA ............... _________
02 mA ............... _________
03 mA ............... _________
04 mA ............... _________
05 mA ............... _________
06 mA ............... _________
07 mA ............... _________
08 mA ............... _________
09 mA ............... _________
10 mA ............... _________
11 mA .............. _________
12 mA .............. _________
13 mA ............... _________
14 mA ............... _________
15 mA ............... _________
16 mA ............... _________
17 mA ............... _________
18 mA ............... _________
19 mA ............... _________
20 mA ............... _________
18
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 19 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-008: TC ultrapure water (Cond)
Select menu:
Parameter setting/Module Cond 7700/TC process medium/Ultrapure water
Menu Display TC process medium
0.20 mS/cm
22.3 °C
TC process medium (Administrator)
EN 27888: Natural waters (0 ... 35°C)
Ultrapure water: traces of impurity
TC correction
Reference temp
Impurity
Ultrapure water
Return OK
NaOH
NaCl
HCl
NH3
Ultrapure water with traces of impurity
• Parameter setting
• Module Cond 7700
• TC process medium
• TC correction: Ultrapure water
Select type of impurity:
• NaOH
Alkaline ultrapure water
• NaCl
Neutral ultrapure water, for conductivity measurement in water processing behind gravel bed filter
• HCl
Acidic ultrapure water, for conductivity measurement behind cation filter
• NH3
Ammoniacal ultrapure water
19
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 20 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-009: Concentration determination
Select menu: Parameter setting/System control/Module Cond/Concentration
Default settings and selection range
Concentration determination
The substance concentration in percent by weight (% by wt) is determined for H
2
SO
4
, HNO
3
, HCl, NaOH, NaCl from the measured conductivity and temperature values.
Conditions for concentration determination
Below you find tables with the concentration ranges of the preset substances.
These tables show the conductivity of the substances versus substance concentration and process temperature:
• For calculation of concentration, the medium to be measured must be a purely binary mixture (e.g. water-hydrochloric acid). Presence of other dissolved substances (e.g. salts) leads to incorrect concentration values.
• In the region of small slopes (e.g. at the range limits) small conductivity changes can correspond to great concentration changes. This may lead to an unsteady display of the concentration value.
• As the concentration value is calculated from the measured conductivity and temperature values, accurate temperature measurement is very important. Therefore, you should make sure that sensor and process medium are in thermal equilibrium.
Messages
You can define limits for warning and failure messages for the concentration value:
Select menu: Parameter setting/System control/Module Cond/Messages
20
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 21 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
Parameter setting: Concentration curves
(Additional function SW 700-009) Default settings and selection range
Note: Function check active
Parameter Default Selection / Range
Concentration
• Medium (“Yes” selected)
No
H2SO4 (0-30%)
USP function
• Reduced limit
• Monitoring
Off
100 %
Off
Yes, No
H2SO4 (0-30 %), H2SO4 (32-84 %), H2SO4 (92-99 %),
HNO3 (0-30 %), HNO3 (35-96 %),
HCl (0-18 %), HCl (22-39 %)
NaOH (0-14 %), NaOH (18-50 %),
NaCl (0-26 %),
Chart (additional function SW 700-009)
Off, On
10 % ... 100%
Off, failure, maintenance request
Sulphuric acid H
2
SO
4 c[% by wt]
21
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 22 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
Nitric acid HNO
3 c[% by wt]
Hydrochloric acid HCl
22
c[% by wt]
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 23 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
Sodium hydroxide solution NaOH c[% by wt]
Table salt solution NaCl c[% by wt]
23
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 24 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-009: Concentration chart (Cond)
Select menu: Parameter setting/System control/Concentration chart
Specifying a concentration solution for conductivity measurement
Concentration chart
To specify the customer-specific solution, 5 concentration values A-E are entered in a matrix together with 5 temperature values 1-5. To do so, first enter the 5 temperature values, then enter the respective conductivity values for each concentration A-E.
These solutions will then be available in addition to the permanently stored standard solutions (select “Chart”).
Menu Display Enter concentration chart
0.020 mS/cm
25.6 °C
System control (Administrator)
Calculation Blocks
Point of measurement
Release of options
Logbook
Concentration chart
Factory setting
Return
Enter values
• Call up parameter setting
• System control
• Select “Concentration chart”
0.020 mS/cm
25.6 °C
Concentration table (Administrator)
Temp 1
Temp 2
Temp 3
Temp 4
Temp 5
+000.0 °C
+005.0 °C
+010.0 °C
+015.0 °C
+020.0 °C
Abort OK
0.020 mS/cm
25.6 °C
Concentration table (Administrator)
Concentration A:
1. Cond at +000.0 °C
2. Cond at +005.0 °C
3. Cond at +010.0 °C
4. Cond at +015.0 °C
5. Cond at +020.0 °C
Abort
05.00 % by wt
0.000 µS/cm
0.000 µS/cm
0.000 µS/cm
0.000 µS/cm
0.000 µS/cm
OK
Enter 5 temperature values
(right/left arrow keys to select position, up/down arrow keys to edit number, confirm with enter.)
Enter values for concentrations A-E for the respective temperatures. The table values must be continuous.
Maxima/minima are not permitted.
Incorrect entries are marked with
The concentration chart is selected as follows:
Parameter setting/Module Cond/Cal preset values/Automatic/Chart.
xx .
24
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 25 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
The chart is built up as 5 x 5 matrix:
Temp 1
Conc. A Conc. B Conc. C Conc. D Conc. E
A1 B1 C1 D1 E1
Temp 2
Temp 3
A2
A3
B2
B3
C2
C3
D2
D3
E2
E3
Temp 4
Temp 5
A4
A5
B4
B5
C4
C5
D4
D5
E4
E5
Conditions for the chart:
• The temperature values must be rising (Temp 1 is the lowest, Temp 5 the highest temperature).
• The concentration values must be rising (Conc. A is the lowest, Conc. E the highest concentration).
• The chart values A1 ... E1, A2 ... E2 etc. must all be rising within the chart or all falling. Points of inflection are not allowed!
The M 700 automatically checks the chart values. Incorrect entries are marked.
25
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 26 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-011: High CO2 compensation
Application-specific additional function for breweries
This additional function simplifies parameter setting since all steps not required for measurement of beer are omitted. It simultaneously acts on all installed O
2 module (module software version 2.2 and higher).
Function principle:
During the filling process it must be ensured that as little oxygen as possible is dissolved in the beer to extent its shelf life.
During oxygen trace measurement the sensor is operated with a very low polarization voltage (-500 mV). This results in low cross-sensitivity to CO
2
.
For a calibration in air, this polarization voltage is too low.
It must be set to -675 mV and afterwards be reduced again to -500 mV for measuring in the trace range.
A sufficiently long waiting time must be observed until the sensor has stabilized.
Opening and closing of valves causes pressure variations in the beer pipes which momentarily falsify the O
2 signal. Therefore the input signal must be attenuated correspondingly to suppress transient interferences.
These processes are automated by the additional function, i.e. all parameters required for the respective program step are set automatically. When the additional function has been activated, the parameter setting steps are displayed as follows:
26
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 27 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
Parameter Default Selection / Range
Input filter
• Pulse suppression
Sensor data
• Measure in
• Sensor type
• Temperature probe **
• Sensor **
• Reference electrode **
• Membrane correction
• Polarization voltage
• Sensocheck
Cal preset values
• Cal saturation
• Cal concentration
• Calibration timer
Pressure correction
• Pressure during meas
• Pressure during cal
Salt correction
• Input
Weak
Liquids
A Standard
NTC 22 k Ω
Without guard
Off
1.00
0500 / 0675 mV
Off
A Standard, B Trace sensor (with guard) * ,
C Trace sensor (without guard), Others
NTC 30 kOhm, NTC 22 kOhm
With guard, Without guard
Off, On
0.50 ... 2.00
For sensor type A, B, C: fixed
Off, Failure, Maintenance request
%AIR mg/l
0000 h
Auto
Auto
Off, Weak, Moderate, Strong, On (000 s ... 999 s)
%Air
µg/l, mg/l, ppb, ppm xxxx h (entry)
Auto, Manual (default value 1013 mbars)
Auto, Manual (default value 1013 mbars)
Salinity
Salinity, Chlorinity, Conductivity
(00.00 g/kg or 0.000 µS/cm, depending on selection)
* Sensocheck not possible for trace sensor with guard, therefore disabled
** Can only be set with “Sensor type Others” selected
27
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 28 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-012:
ISFET for ISM module pH 2700i(X)
pH measurement with InPro 3300 ISFET sensor pH 2700i(X) jumper
Note:
Each time a new sensor is connected, an ISFET zero point adjustment must be performed.
After that, you should perform one of the following calibration methods:
• Calimatic: automatic calibration
• Manual: entry of buffer values
• Data entry: premeasured electrodes
28
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 29 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
Parameter settings: New
Modul pH 2700i (Administrator)
7.00 pH
20.1 °C
Input filter
Sensor data
Cal preset values
TC process medium
ORP/rH value
Delta function
Return Block
7.00 pH
20.1 °C
Sensor data (Administrator)
Sensor type ISFET
Temperature detection
Sensoface
Sensor monitoring details
(Pt100)
On Off
Abort OK
7.00 pH
20.1 °C
Sensor monitoring details (Administrator)
Slope
Zero
Sensocheck ref el
Sensocheck glass el
Response time
ISFET leak current
Return
(Auto)
(Auto)
(Auto)
(Auto)
(Auto)
(Auto)
Parameter setting/ module pH 2700i / Sensor data
Sensor data are preset depending on the sensor type.
Gray display lines cannot be edited.
Sensor data
Option SW 700-012 allows the setting of “ISFET” as Sensor type.
Sensor monitoring details
Monitoring of ISFET leak current is available.
Sensor monitor pH input
ORP input
RTD
Temperature
Impedance ref (25°C)
ISFET characteristic
Return
7.00 pH
22.3 °C
43 mV
-109 mV
1.100 k Ω
25.6 °C
322.8 k
Ω
0,81 mA/V
7.00pH
24.1°C
Sensor network diagramm PH
1 - Slope
2 - Zero point
3 - Ref mpedance
4 - ISFET Leak current
5 - Response time
6 - Calibration timer
7 - Sensor wear
Return
Diagnostics: New
Sensor monitor
ISFET characteristic.
Sensor network diagramm pH
New: ISFET leak current.
29
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 30 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
Inserting the SmartMedia card
To release an additional function via TAN, see Pg 6, Release of options
Inserting and replacing the SmartMedia card
The SmartMedia card may be inserted or replaced with the power supply switched on. Protect against electrostatic discharge! When closing the device, make sure that the sealing is properly seated and clean.
Warning!
Do not touch the terminal compartment, there may be dangerous contact voltages!
1. To open the analyzer
• Loosen the 4 front screws
• Open the M 700 FRONT at its right side
(pivot hinge inside at the left)
• The slit for inserting the
SmartMedia card is located at the inner side of the M 700 FRONT module
2. To insert the SmartMedia card
• Remove SmartMedia card from its package without touching the contact surface
• Insert card in the slit at the inner side of the
M 700 FRONT module
Inserting the
SmartMedia card:
The label must be facing you.
30
3. To remove the SmartMedia card
• Please call up the Maintenance menu to avoid data loss.
• “Close memory card” to terminate software access to the SmartMedia card. Now the card can be taken out.
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 31 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SmartMedia card: Usage
Use as memory card in combination with additional functions.
Additional functions must be ordered separately (release via TAN).
A SmartMedia card is provided with the following additional functions:
• Software update (SW 700-106, see Pg 43)
• Additional function SW 700-1xx (description Pg 34 et seq.)
Using commercially available SmartMedia cards
Commercially available SmartMedia cards can be used as additional memory card.
The following types of cards are supported: 8 MB, 16 MB, 32 MB, 64 MB,
128 MB. Externally produced files, such as from a digital camera, are tolerated. Long file names can be read. The M 700 generates file names in the
8.3 format (8 characters file name, 3 characters program-specific file name extension).
Formatting a commercial SmartMedia card
Some devices (e.g. digital cameras, scanners) cause a formatting of the
SmartMedia card which does not correspond to the SSFDC specification or the SmartMedia Interface Library (SMIL). The manufacturer therefore recommends to format a commercial SmartMedia card as M 700 memory card prior to first use.
Menu Display Format SmartMedia card
0.003 mS/cm
20.4 °C
Memory card (Administrator)
Record logbook On
Record recorder
Decimal separator
Card full
On
Point
Record stop
Off
Off
Comma
Data saving
Format card
Return
Formatting
• Insert SmartMedia card
• Open menu selection
• Parameter setting, Admin. level
• Enter passcode
• System control: Memory card
(The “Memory card” function is only available with the SmartMedia
Card inserted!)
• Format card
31
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 32 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
Loading /saving configuration
System control/Memory card/Copy configuration.
Additional functions must be ordered separately (release via TAN).
Saving / loading the complete device configuration
System control/Memory card/Copy configuration.
“Save“ configuration means that the complete device configuration (except the passcodes) is written on the memory card.
“Load“ configuration means that the complete device configuration is read from the memory card and pogrammmed in the M 700.
BACKUP file generated on SmartMedia card: \BACKUP\BACKUP01.PAR
Copy configuration from one M 700 to further M 700 units
(Copying parameter sets for similarly equipped M 700)
Prerequisite:
The devices must have the same hardware.The module equipment must be absolutely identical – the same modules in the same slots!
Options:
All required options must be enabled on the “master device”, this is not required for the“slave devices”.
It is the parameters of the options that are transferred, not the option itself.
When an option will be enabled on a “slave device” at a later point in time, the parameters of this option will be preset corresponding to the “master device”.
1 Write device configuration of configured device on SmartMedia card:
Parameter setting/System control/Copy configuration/Save.
2 Call up Maintenance menu. Select “Close memory card”.
3 Remove SmartMedia card.
Now you can transfer the device configuration to further identically equipped devices.
4 To do so, insert the SmartMedia card containing the configuration in the next device to be configured. Select
Parameter setting/System control/Copy configuration/Load.
5 Call up Maintenance menu. Select “Close memory card”.
6 Remove SmartMedia card.
32
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 33 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SmartMedia card: Format card
Parameter setting/System control/Format card
Menu Display
Format card
(Generate memory card)
0.003 mS/cm
20.4 °C
Software update (Administrator)
Updating will change the device characteristics. Perform verification as appropriate!
Update firmware
Save firmware
Format card
Return
0.003 mS/cm
20.4 °C
Format card (Administrator)
Attention: All data on the memory card are deleted!
Return Format
Format card
• Insert SmartMedia card (Pg 30)
• Open menu selection
• Parameter setting, Admin. level
• Enter passcode
• System control: Format card
Caution!
Formatting an update card generates a memory card.
Refer to “SmartMedia card”, Pg 34.
This process is irreversible! Double warning messages proect against faulty operation.
Note:
The SmartMedia card must always be formatted in the M 700.
Correct functioning of a SmartMedia card with different formatting cannot be guaranteed.
File structure of a memory card (example):
Folder
BACKUP
RECORDER
Typ. file name
BACKUP01.PAR
LOGBOOK L_YYMM00.TXT
PARASET SET 1
R_YYMMDD.TXT
Remark
Parameter set (Backup)
Logbook file,
YY=year, MM=month
Parameter set
Recorder entry,
YY=year, MM=month, DD=day
33
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 34 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SmartMedia card: Memory card
Parameter setting/System control/Memory card
Menu Display
34
Using the memory card
0.003 mS/cm
20.4 °C
Software update (Administrator)
Updating will change the device characteristics. Perform verification as appropriate!
Update firmware
Save firmware
Format card
Return
System control (Administrator)
Memory card
Copy configuration
Parameter sets
Function control matrix
Time/date
Point of measurement
0.003 mS/cm
25.6 °C
Return
0.003 mS/cm
25.6 °C
Memory card (Administrator)
Record logbook On
Record recorder On
Decimal separator
Card full
Point
Record stop
Format card
Off
Off
Comma
Return
Using the memory card
• Insert SmartMedia card (Pg 30)
• Open menu selection
• Parameter setting, Admin. level
• Enter passcode.
• System control: Memory card
With SmartMedia card inserted, the display shown on the left appears
(The “Memory card” line is displayed only if there really is a memory card in the slot).
• Select “Memory card”, confirm with enter.
The menu is self-explanatory.
Behavior when the memory card is full:
Continuous recording (as with a flight recorder) or Stop (card replacement).
0.003 mS/cm
25.6 °C
Copy configuration (Administrator)
Identical module equipment required for transfer.
Configuration
Return
Save
Load
Save
Copy configuration
• Save: saves all data on the memory card
• Load: Overwrites all device data with the data from the memory card
Caution!
“Close” memory card before removing it (Maintenance menu)
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 35 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-102: Loadable parameter sets
Parameter setting/System control/Parameter sets
Menu Display
Load parameter set from
SmartMedia card
Parameter sets (Administrator)
0.003 mS/cm
25.1 °C
Select with softkey
Settings in menu
Softkey usage
Parameter sets
Save parameter set A
Load parameter set A
A,B (internal)
1,2 (card)
1,3 ,3 (card)
1,2,3,4 (card)
Abort
1,2,3,4,5 (card)
OK
Parameter sets (Administrator)
0.003 mS/cm
25.0 °C
Select with softkey
Settings in menu
Softkey usage
1,2,3,4 (card) Parameter sets
Save parameter set A
Load parameter set A
Return
Loading a parameter set from
SmartMedia card
• Call up parameter setting
• System control
• Call up “Parameter sets” (Fig)
The M 700 provides 2 complete parameter sets (A, B).
5 parameter sets can be stored on the SmartMedia card. One of those can be saved as parameter set A to the M 700:
SmartMedia card 0.003 mS/cm
24.9 °C
Load parameter set A (Administrator)
NOTE: Old parameter set will be overwritten!
B (internal)
1 (card)
Load from 2 (card)
3 (card)
Abort
Set 1 Set 2 Set 3
Set A
Set B
(internal only)
0.003
25.0
mS/cm
°C
Time 11:49 Favorites menu
• Select parameter set to be loaded
Activated parameter set is displayed in measuring mode.
Note: Remote switching between A and B is possible via the OK2 input.
35
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 36 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-102: Loadable parameter sets
Parameter setting/System control/Parameter sets
Menu Display
Save parameter set on
SmartMedia card
Parameter sets (Administrator)
0.003 mS/cm
25.1 °C
Select with softkey
Settings in menu
Softkey usage
Parameter sets
Save parameter set A
Load parameter set A
Abort
A,B (internal)
1,2 (card)
1,2,3 (card)
1,2,3,4 (card)
1,2,3,4,5 (card)
Parameter sets (Administrator)
0.002 mS/cm
24.2 °C
Select with softkey
Settings in menu
Softkey usage
1,2,3,4 (card) Parameter sets
Save parameter set A
Load parameter set A
Return
To save a parameter set on
SmartMedia card
• Call up parameter setting
• System control
• Call up “Parameter sets” (Fig)
The M 700 provides 2 complete parameter sets (A, B).
Up to 5 parameter sets can be loaded to the SmartMedia card.
To do so, a parameter set
(1, 2, 3, 4, or 5) of the SmartMedia card is overwritten by parameter set
A of the M 700.
0.002 mS/cm
24.2 °C
Save parameter set A (Administrator)
NOTE: Old parameter set will be overwritten!
B (internal)
1 (card)
Save to 2 (card)
3 (card)
Abort
4 (card)
5 (card)
• Select parameter set on SmartMedia card
Set A
Set B
(internal only)
Set 1 Set 2 Set 3
SmartMedia card
36
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 37 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-103: Measurement recorder
Select menu: Parameter setting/FRONT M 700/Measurement display
The last 200 data records are stored in the M 700 and are represented graphically – like on a recorder. All further data can be stored on the
SmartMedia card and be evaluated by a computer.
2 process variables are represented. You can configure:
• The process variable to be displayed
• Start and end value for the variable to be recorded
• Time base (recording interval, selectable from 10 s to 10 h)
In addition, the time axis can be stretched by factor 10 using the “Zoom function”. In the event of quick changes, the zoom function is automatically activated. It always starts several pixels before the event. This allows detailed investigation of signal instabilities.
With “Min/Max displays” activated, the respectively measured peak values
(gray) are shown in addition to the measured value (average).
0.002 mS/cm 0.004 mS/cm 20°C 25°C
Graphic range
Start and end value of process variable
S/cm
07/24/02 11:32, Status: Measurement
0.003mS/cm, 23.9°C
1 2
Info
°C
10x
Events (NAMUR) (see “Info”)
Maintenance request
Function check
Failure
Min/Max display
The minimum and maximum values within the recording interval are shown as a gray line next to the (averaged) measured value.
Fig. 1: Min/Max off
Fig. 2: Min/Max display
37
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 38 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-103: Measurement recorder
Select menu: Parameter setting/FRONT M 700/Measurement display
Additional function SW 700-103: Measurement recorder
The measurement recorder saves all entries in a file. The last 200 entries can be displayed on the M 700. A new file is generated for each day. The date is encoded in the file name.
Example for a file generated on SmartMedia card:
\RECORDER\R_YYMMDD.TXT
Recorder data of YYMMDD
(YY = year, MM = month, DD = day)
The data is recorded as ASCII file with the extension .TXT.
The individual columns are separated by tabs. This makes the file readable with word processing or spreadsheet programs (e.g. Microsoft Excel). Each time the memory card is inserted in the slot, a “Device Info” consisting of
Model number, BASE serial number, and tag number. Thus, a memory card can also be used to collect the measurement recorder data of several devices.
Example:
M 700 - Measurement recorder
Time stamp CH Left
<< M 700 - Serial 002004 [DSE KL_007] >>
21.10.02 00:09:15
21.10.02 00:19:15
21.10.02 00:29:15
21.10.02 00:39:15
21.10.02 00:49:15
21.10.02 00:59:15
1015
1015
1015
1015
1015
1015 mbar mbar mbar mbar mbar mbar
2.8
2.9
2.8
2.9
2.9
3.0
CH Right
°C
°C
°C
°C
°C
°C
M-req Fct.
0
0
0
0
0
0
0
0
0
0
H
0
Fail
0
0
0
0
H
0
Slow
0
0
0
0
0
0
38
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 39 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-103: Measurement recorder
Select menu: Parameter setting/FRONT M 700/Measurement display
Meaning of the entries in the recorder file:
Time stamp Time stamp of recorder entry
CH Left Left channel of recorder with measured value and unit
CH Right
M-req.
Fct.
Fail
Slow
Right channel of recorder with measured value and unit
NAMUR signal* “Maintenance request”
NAMUR signal* “Function check”
NAMUR signal* “Failure”
Zoom function active (Slow motion)
* Definition:
NAMUR = German committee for measurement and control standards in the chemical industry
39
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 40 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-104 : Extended logbook
Parameter setting/System control/Logbook
Additional function SW 700-104: Extended logbook
The extended logbook saves all entries in a file. The last 50 entries can be displayed on the M 700. A new file is generated for each month. The date is encoded in the file name.
Example for a file generated on SmartMedia card:
\LOGBOOK\L_YYMM00.TXT
Recorder data of YYMM
(YY = year, MM = month)
The data is recorded as ASCII file with the extension .TXT.
The individual columns are separated by tabs. This makes the file readable with word processing or spreadsheet programs (e.g. Microsoft Excel). Each time the memory card is inserted in the slot, a “Device Info” consisting of
Model number, BASE serial number, and tag number is written. Thus, a memory card can also be used to collect the logbook data of several devices.
Example:
M 700 - Logbook
Time Stamp Status Message
<< M 700 - Serial 0002004 [DSE KL_001] >>
21.11.02
19:08:43
22.11.02
06:02:01
22.11.02
06:09:27
22.11.02
06:09:36
23.11.02
16:45:07 (x)
Power supply Off
Power supply On
Diagnostics active
Measurement active
Fail current I2 > 20 mA
23.11.02
18:43:11
23.11.02
18:47:38
23.11.02
18:47:38 ( )
:
:
Parameter setting active
Measurement active
Fail current I2 > 20 mA
Time stamp:
Status
Message
40
Time stamp of logbook entry
(x) - Message activated
( ) - Message deactivated
Message text (in selected operator language)
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 41 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-106: Software update
For a software update (additional function SW700-106), the manufacturer supplies a specially formatted SmartMedia card. The M 700 replaces its own firmware (operating program) by the new version (“Update”).
Caution!
During a software update the M 700 is not operable!
After a software update you should check the configuration.
UP
This icon indicates that a SmartMedia card is inserted in the slot.
The card allows storing of current device software on the card as well as loading of new software into the M 700.
1. Save the firmware currently installed in your M 700 (Pg 42).
2. Load the software update as described on Pg 43.
Note:
A memory card can be generated by formatting an update card (irreversible!).
See Page 31.
41
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 42 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SmartMedia card: Save firmware
Parameter setting/System control/Software update/Save firmware
Menu Display
Save firmware on software update card
UP
0.003 mS/cm
20.4 °C
Software update (Administrator)
Updating will change the device characteristics. Perform verification as appropriate!
Update firmware
Save firmware
Format card
Return
UP
0.003 mS/cm
20.4 °C
Save firmware (Administrator)
UP
Select a free memory slot or overwrite a stored firmware version
Save firmware
• Insert SmartMedia card
• Open menu selection
• Parameter setting, Admin. level
• Enter passcode
• System control: Software update
Select a free memory slot on the card:
Slot 2: SZ34Z10000/0-V1.4
Return Save
UP
0.003 mS/cm
20.4 °C
Save firmware (Administrator)
UP
• Select slot with key
• Select free slot with arrow key.
Slot
5: <empty>
6: <empty>
7: <empty>
Return OK
UP
0.003 mS/cm
20.4 °C
Save firmware (Administrator)
Select a free memory slot
UP or overwrite a stored firmware version
• Confirm slot with “OK”.
Slot 3: <empty>
Return
UP Save successful
Save
Start with “Save” softkey.
Confirm finish message
(with “OK” or enter).
Remove the SmartMedia card.
Close the front door.
42
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 43 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW 700-106: Software update
Parameter setting/System control/Software update/Load firmware
Menu Display
Software update
(“Load firmware”)
0.003 mS/cm
25.6 °C
Release of options (Administrator)
102 Parameter sets
103 Measurement recorder
104 Logbook
106 Software update
107 AuditTrail
Inactive
Inactive
Inactive
Active
Inactive
Software update
• Insert SmartMedia card (Pg 30)
• Open menu selection
• Parameter setting, Admin. level
• Enter passcode
• Select System control
Return
0.003 mS/cm
20.4 °C
Software update (Administrator)
Updating will change the device characteristics. Perform verification as appropriate!
Update firmware
Save firmware
Format card
Return
UP
0.003 mS/cm
20.4 °C
Update firmware (Administrator)
UP
Select the memory slot for
( the chosen firmware indicates the existing version)
Slot 1: SZ34Z10000/0-V1.0
Abort OK
UP
0.003 mS/cm
20.4 °C
Update firmware (Administrator)
UP
Select the memory slot for
( the chosen firmware indicates the existing version)
1: Select Release of options
(Software update SW 700-106)
Set option to “active”. Enter the
TAN at the prompt. The option is available after the TAN has been entered.
2: Select Software update
Check whether your unit really requires a software update!
To read the current software version, select:
• Diagnostics
• Device description
• FRONT M 700
Perform update:
• Parameter setting
• System control
• Software update
• Select slot
• Confirm slot with “OK”.
• Press “Load firmware” softkey to start software update.
Slot 1: SZ34Z10000/0-V1.0
Return Load firmware
43
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 44 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 45 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
AuditTrail and signature
Electronic data recording and protection according to FDA 21 CFR Part 11
Additional function SW 700-107
Provided by: TAN and AuditTrail card
Contents
AuditTrail....................................................................................................46
Signature....................................................................................................47
Activating the AuditTrail .............................................................................48
Signature entry...........................................................................................49
AuditTrail: Parameter setting.......................................................................50
AuditTrail: User management .....................................................................51
AuditTrail: Recording ..................................................................................52
AuditTrail log ..............................................................................................53
45
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 46 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
AuditTrail
Function description
Electronic data recording and protection according to
FDA 21 CFR Part 11
Pharmaceutical applications not only require a hygienic design, but also the protection and complete recording of data (FDA regulation: 21 CFR Part 11).
The M 700 with stainless steel enclosure and SmartMedia card meets these demands. The FDA regulation basically consists of two parts: “Audit Trail” and “Electronic Signature”. Both parts interact directly: For every change of parameters, the device records what has been changed and who has changed it.
Functions like calibration (adjustment) often may only be performed after written approval! This approval is given by an electronic signature.
AuditTrail card
The AuditTrail card is a specially prepared SmartMedia card. The M 700 records the AuditTrail log on an “AuditTrail card” only. Readability from PC is not affected.
Note:
Only specially coded SmartMedia cards supplied from Mettler Toledo can be used as AuditTrail cards.
As delivered, the AuditTrail card includes a PC program which makes the recorded AuditTrail data readable. When log coding has been activated, it also allows decoding of the data. The program supports the data export to other programs, e.g. Microsoft Excel.
46
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 47 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
Signature
Function description
Signature
A signature consists of two entries, the user name and the corresponding passcode. For each change at the device (calibration/adjustment, parameter setting, maintenance, if required also diagnostics), the M 700 records in the
“AuditTrail Log” and in addition in the logbook who has performed the change. For that purpose, access authorizations are introduced.
Access authorization
Every access authorization consists of a signature:
• User name and
• Passcode.
The administrator and the user enter the access data in the user management menu. The administrator enters the user name (short name) and the user adds his/her personal passcode.
Each time he/she calls up a protected menu, the user must enter his/her name and passcode. The M 700 only gives access to the function if the access codes are valid. The recordings (AuditTrail log, logbook) on the
AuditTrail card include the user names so that all actions can be clearly traced back.
Coded storage
The entries in the AuditTrail log can be coded using an algorithm and protected with checksums so that a special evaluation program can prove the authenticity of the data records on a PC.
47
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 48 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
Activating the AuditTrail
Select menu: Parameter setting/System control/Release of options
Note: The TAN for releasing an additional function is only valid for the device with the corresponding serial number
Menu Display Activate additional function
48
Menu selection
7.00 pH
25.6 °C
Select:
Return to meas
[enter]
Lingua
25.6 °C
0.003 mS/cm
Parameter setting
Viewing level
Operator level
Admin. level
(All Data) view
(Operation Data) opl
(All Data) adm
Menu selection
Call up parameter setting.
From the measuring mode:
Press menu key to select menu.
Select parameter setting using arrow keys, confirm with enter.
Parameter setting
Select Administrator level using arrow keys, confirm with enter.
Enter passcode and confirm
(Passcode as delivered: 1989).
Return
16.9 %Air
25.6 °C
Parameter setting (Administrator)
System control
FRONT M 700-011
BASE M 700-021
Module O2 4700
Module PH 2700
Module Cond Ind 7700
Return
0.003 mS/cm
25.6 °C
Release of options (Administrator)
102 Parameter sets
103 Measurement recorder
104 Logbook
106 Software update
107 AuditTrail
Active
Inactive
Inactive
Inactive
Active
Inactive
Abort
Select System control using arrow keys, confirm with enter.
Then select Release of options using arrow keys, confirm with
enter.
Release of options
Select the additional function to be released.
Set option to “active”. Enter the
TAN at the prompt. (Note: The TAN is only valid for the device with the corresponding serial number, see previous page.) The option is available after the TAN has been entered.
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 49 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
Signature entry
(User name, passcode)
Signature entry (user name, passcode) to open a menu
To protect the data against unauthorized query or modification, access data must be entered before a main menu from the menu selection is opened.
Two-stage security is provided by input of user name and passcode.
Menu selection
7.00 pH
25.6 °C
Enter signature:
Select:
User: ADMIN - - -
Passcode:
• • •
[enter]
Return to meas Lingua
The passcode is not visible. It is only displayed as 4 dots. The administrator can delete each user’s passcode (with entry in
AuditTrail log). The user is identified by his/her access data.
After entry of the valid signature all menus are available with the defined user rights. Only in measuring mode, the rights will be erased again. This allows accessing several menus without having to enter the access data each time.
When an invalid combination of user name and passcode is entered, access is denied with the text “Wrong passcode”. The entry “Wrong passcode” with the entered (wrong) user ID is recorded in the AuditTrail log.
Default setting for signature
Signature
User: ADMIN
Passcode: 1989
49
SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 50 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
AuditTrail: Parameter setting
Adapt function: Parameter setting/System control/AuditTrail
Parameter setting
In the System Control menu the Administrator can adjust the AuditTrail system to the requirements of his company. For example, he/she can select the protection against unauthorized query and the coding of the data in the
AuditTrail log.
AuditTrail (Administrator)
User management
AutoComplete
Access diag menu
Record AuditTrail
Coding Log
On
On
On
7.20 pH
25.6 °C
Off
Free
Off
Off
Return
• “AutoComplete” is an input assistance for the user name as known from some
Windows programs.
• Access to the Diagnostics menu can be
“free” or require input of a “Signature” according to the access authorization granted the User Management menu.
• For the recording on SmartMedia card, a coding can be activated so that the data cannot be read or queried without corresponding programs.
User management
An administrator and up to 5 users can be entered in the M 700. Each user has a user name (max. 8 characters) and a passcode (4 characters). Each user can be granted access and usage rights for each main menu level. The combination of user name and passcode is the user’s “Electronic Signature”.
Every logged modification is signed and can be clearly traced back. Different access rights for each menu can be defined for each user.
50
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:09 Uhr Seite 51 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
AuditTrail: User management
Create, edit, erase user data
Menu Display AuditTrail: User management
Menu selection
Select:
Return to meas
7.0 pH
17.1 °C
[enter]
Lingua
7.20 pH
25.6 °C
User management (Administrator)
Admin: ADMIN
User 1: BERT
User 2: ERNIE
User 3: KERMIT
User 4:
User 5:
(adj, maint, adm, diag)
(cal; maint; opl, diag)
(view)
(view, diag)
()
()
Return
7.20 pH
25.6 °C
User 1: Setup (Administrator)
User name
Passcode cal level maint level par level diag level
BERT - - - -
••••
Check
Maintenance
Operator
Diagnostics
Return
Menu selection
From the measuring mode:
Press menu key to select menu.
Select parameter setting using arrow keys, confirm with enter.
Signature for first access:
User: ADMIN
Passcode as delivered: 1989
Select:
Parameter setting/System control/
AuditTrail/User management.
Access data and rights can be created for the administrator and up to 5 users.
The administrator has always administrator rights for parameter setting.
The user enters his/her passcode after being asked to do so by the administrator. The passcode is only displayed as a series of 4 dots.
Access rights for the main menu levels:
• Calibration:
• Maintenance:
[ No access | Check | Adjustment ]
[ No access | Maintenance ]
• Parameter setting: [ No access | Viewing | Operator | Administrator ]
• Diagnostics: [ No access | Diagnostics ]
51
SW700_ENG_Version6.QXD 01.08.2005 14:09 Uhr Seite 52 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
Record AuditTrail
Functions
AutoComplete
The “AutoComplete” function facilitates the input of the user name. You only have to enter as many letters as the M 700 requires to identify the name.
This function can be switched off in the AuditTrail menu because the automatic display of the user name reduces system security, which then might not be sufficient for FDA documentation.
Access to diag menu
To protect all data against unauthorized query, it is also possible to protect the diagnostics level with a passcode (this will also affect the Favorites menu). As delivered, access to the diagnostics level is free.
Record AuditTrail
The AuditTrail log is an additional - protected if required - recording of all data on the AuditTrail card (SmartMedia card). The log is recorded in the currently selected language.
The data can be coded (to protect against unauthorized query) and secured with a checksum (to prove the authenticity of the data). The coding can be switched on or off in the AuditTrail menu of the Administrator level.
A four-digit line numbering (0000 ... 9999 P 0000 ..., cyclic) ensures uninterrupted recording. The AuditTrail log cannot be deleted by the M 700.
A buffer storage is provided to store the recorded data while the AuditTrail card is being replaced, for example. An error message “Fail AuditTrail card” is generated when the buffer is full. The oldest entries in the buffer will be recorded. They are then missing in the log (the corresponding line numbers will be missing).
When an empty AuditTrail card is inserted, the buffer storage and the current entries are transferred to the empty card.
52
M 700(X)
SW700_ENG_Version6.QXD 01.08.2005 14:09 Uhr Seite 53 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
Record AuditTrail
Parameters
The AuditTrail log cannot be queried by the M 700. It can only be evaluated using the suitable PC program. According to FDA 21 CFR Part 11 the user must ensure the transport of the data from the M 700 to evaluation and further processing or archiving. Furthermore, the user must ensure that the log data cannot be deleted from the memory card. The PC evaluation program is able to decode the data and verify the security parameters and thus to prove the authenticity of the data on the card. It allows printing the encoded data and exporting them to other PC applications.
Logbook
The logbook is provided in addition. In AuditTrail systems it runs simultaneously with the audit trail log in contrast to which it can be queried provided that the diagnostics menu is accessible. The administrator is permitted to delete the logbook data since all security-relevant data are stored in the
AuditTrail log.
Recordings in the AuditTrail log
Caution!
Data loss (even total destruction) if the card is not closed before it is pulled out. (Maintenance menu, “Close card”)
The following entries and messages are recorded:
• Measuring point, device description, module equipment and serial numbers
• Menu system log-ins
• Movements within menus (menu headlines)
• Modified parameters at the press of enter
• Calibration record at the end of a calibration or adjustment incl. user name, sensor type and serial number
• Messages such as failure and maintenance request
53
SW700_ENG_Version6.QXD 01.08.2005 14:09 Uhr Seite 54 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
AuditTrail log
Measurement point
FRONT M 700-011
BASE M 700-021 pH 2700
O2 4700
BIGBLUE vessel
0006123
0006458
0007221
0006045
6653 07/01/04 10:43:02Login: Admin (ADMIN)
6654 07.01.04
10:43:02Login: > Parameter setting (Administrator)
6655 07.01.04
10:43:02Login: > [1] Module pH 2700 (Administrator)
6656 07.01.04
10:43:02Login: < Parameter setting (Administrator)
6657 07.01.04
10:43:02Login: < Menu selection
6658 07.01.04
10:43:02Login: Admin
6659 07.01.04
10:43:02Login: > Calibration
6660 07.01.04
10:43:02Login: > [1] Module pH 2700
6661 07.01.04
10:43:02Login: > [1] Calimatic
6662 07.01.04
10:43:02Login: Warn Identical Buf
6663 07.01.04
10:43:02Login: > [1] Calimatic
6664 07.01.04
10:43:02Login: [1] Cal record
6665
6666
6667
6668
6669
Last calibration: 20.11.03 10:44
User: ADMIN
Sensor type: El 204
Serial number: 0815
Cal mode: Calimatic
6670
6671
6672
6673
Zero: 7.02 pH
Slope: 58.0 mV/pH
Impedance glass (25°C): 825.9 MOhm
Impedance ref (25°C): 119.4 kOhm
6674 0
6675
6676
6677
1st nominal buffer value: 7.00 pH
Desired value: 6.97 pH
Actual value: 7.10 pH
Delta value: 0.13 pH
6678 0
6679 0
Electrode potential: 1 mV
Cal temperature: 25.0 °C
6680 0 Response time: 19 s
6681 07.01.04
10:43:02Login: < [1] Module pH 2700
6682 07.01.04
10:43:02Login: < Calibration
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
Typical AuditTrail log
(gray: logged measurement deviations)
SW700_ENG_Version6.QXD 01.08.2005 14:09 Uhr Seite 55 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
SW700_ENG_Version6.QXD 01.08.2005 14:09 Uhr Seite 56 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S
Additional functions
Provided by: device-specific TAN
To release: Parameter setting / System control / Release of options
Passcode Administrator level 1989 (new: ...............)
52121198 KI recorder ..........................................................7
52121199 Buffer sets to be entered (pH) ...........................13
52121200 Adaptive cal timer (pH)......................................13
52121201 ServiceScope (pH)..............................................14
52121202 Tolerance band recorder (pH).............................15
52121203 Current characteristic definable .........................17
52121204 TC ultrapure water (Cond) ................................19
52121205 Concentration determination (Cond) .................20
52121250 High CO
2 compensation (O
2
) ............................26
52121274 ISFET for ISM module pH 2700i(X).....................28
Additional functions
Provided by: device-specific TAN
Inserting the SmartMedia card..............................................30
Removing the SmartMedia card............................................30
Usage ...................................................................................31
52121192 5 loadable parameter sets .................................35
52121193 Measurement recorder ......................................37
52121194 Extended logbook .............................................40
52121195 Software update ...............................................43
52121241 AuditTrail (FDA 21 CFR Part 11).........................45
advertisement
Key Features
- Expandable modular process analysis system
- Device-specific additional functions to enhance capabilities
- Data logging for comprehensive process monitoring
- Tolerance band monitoring for quality control
- Concentration determination for precise measurements
- High CO2 compensation for accurate O2 measurement
- ISFET for ISM module pH 2700i(X) for reliable pH measurement
- 5 loadable parameter sets for flexible configuration
- Measurement recorder for data analysis and trending
- Extended logbook for detailed record-keeping
Related manuals
Frequently Answers and Questions
How do I activate an additional function?
What is the benefit of using the Tolerance band recorder?
Can I use the M 700(X) to measure pH and conductivity simultaneously?
advertisement
Table of contents
- 3 Additional functions
- 2 Warranty
- 2 Registered trademarks
- 4 Latest product information
- 5 How to order an additional function
- 6 Activate additional function
- 7 Provided by: TAN
- 7 SW 700-001: KI recorder
- 13 SW 700-002: Buffer set to be entered (pH)
- 14 SW 700-004: ServiceScope (pH)
- 15 SW 700-005: Tolerance band recorder (pH)
- 17 SW 700-006: Current characteristic definable
- 19 SW 700-008: TC ultrapure water (Cond)
- 20 SW 700-009: Concentration determination
- 26 SW 700-011: High CO2 compensation
- 28 SW 700-012*: ISFET for ISM module pH 2700i(X)
- 30 Provided by: TAN. SmartMedia card required.
- 30 SmartMedia card
- 30 Inserting the SmartMedia card
- 31 SmartMedia card: Usage
- 32 Loading /saving configuration
- 33 SmartMedia card: Format card
- 34 SmartMedia card: Memory card
- 35 SW 700-102: Loadable parameter sets
- 37 SW 700-103: Measurement recorder
- 40 SW 700-104: Extended logbook
- 41 SW 700-106: Software update
- 42 SmartMedia card: Save firmware
- 43 SmartMedia card: Load firmware
- 45 Provided by: TAN and AuditTrail card
- 46 AuditTrail
- 47 Signature
- 48 Activating the AuditTrail
- 49 Signature entry
- 50 AuditTrail: Parameter setting
- 51 AuditTrail: User management
- 52 Record AuditTrail
- 53 AuditTrail log
- 56 Additional functions