Mettler Toledo transmitter M 700(X) - additional functions Operating Instruction

Add to My manuals
56 Pages

advertisement

Mettler Toledo transmitter M 700(X) - additional functions Operating Instruction | Manualzz

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 1 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW

M 700(X)

Additional functions

O pt ion retrofitt ab le

Latest product information: www.mtpro.com

Order number: 52 121 241

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 2 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW

Warranty

Defects occurring within 1 year from delivery date shall be remedied free of charge at our plant

(carriage and insurance paid by sender).

©2005 Subject to change without notice

Return of products under warranty

Please contact your local Mettler-Toledo representative before returning a defective device. Ship the cleaned device to the address you have been given. If the device has been in contact with process fluids, it must be decontaminated/disinfected before shipment. In that case, please attach a corresponding certificate, for the health and safety of our service personnel.

Disposal

Please observe the applicable local or national regulations concerning the disposal of

“waste electrical and electronic equipment”.

Registered trademarks

The following registered trademarks are used in this instruction manual without further marking

CalCheck

Calimatic

Sensocheck

Sensoface

ServiceScope

VariPower

SMARTMEDIA ® is a registered trademark of Toshiba Corp., Japan

InPro ® is a registered trademark of Mettler Toledo GmbH, Switzerland

Mettler-Toledo GmbH, Process Analytics, Industrie Nord,

CH-8902 Urdorf, Tel. +41 (44) 736 22 11 Fax +41 (44) 736 26 36

Subject to technical changes. Mettler-Toledo GmbH, 11/05.

Printed in Germany.

TE-201.011-MTE06 071105

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 3 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW

Additional functions

for the M 700(X) modular measuring system

Additional functions

Latest product information ...........................................................................4

How to order an additional function ............................................................5

Information on TAN release and SmartMedia cards

Activating an additional function (Option release) ........................................6

SmartMedia card ........................................................................................30

Provided by: TAN

52121198 SW 700-001 KI recorder .........................................................7

52121199 SW 700-002 Buffer sets to be entered (pH) ..........................13

52121201 SW 700-004 ServiceScope (pH).............................................14

52121202 SW 700-005 Tolerance band recorder (pH)............................15

52121203 SW 700-006 Current characteristic definable ........................17

52121204 SW 700-008 TC ultrapure water (Cond) ...............................19

52121205 SW 700-009 Concentration determination (Cond) ................20

52121250 SW 700-011 High CO

2 compensation (O

2

) ...........................26

52121274 SW 700-012 ISFET for ISM module pH 2700i(X)....................28

Provided by: TAN. SmartMedia card required.

52121192 SW 700-102 5 loadable parameter sets ................................35

52121193 SW 700-103 Measurement recorder .....................................37

52121194 SW 700-104 Extended logbook ............................................40

52121195 SW 700-106 Software update ..............................................41

52121196 SW 700-107 AuditTrail and Signature ...................................45

3

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 4 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW

Latest product information

Additional functions

The M 700 is an expandable modular process analysis system.

For latest product information, please refer to: www.mtpro.com

Additional functions

A current version of the respective function description is available as pdf for download.

4

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 5 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW

How to order an additional function

Device-specific TAN (transaction number)

Additional functions expand the device capabilities.

The additional functions are device-specific. When ordering an additional function, you therefore have to specify the serial number and hardware version of your M 700 FRONT in addition to the respective order number.

The manufacturer then supplies a TAN (transaction number) to release the additional function.

Serial number of M 700 FRONT

Menu Display Serial number FRONT M 700

Menu selection

0.002 mS/cm

23.7 °C

Menu selection

Call up diagnostics.

From the measuring mode:

Press menu key to select menu.

Select diagnostics using arrow keys, confirm with enter.

Select:

Return to meas

[enter]

Lingua

0.003 mS/cm

23.7 °C

Diagnostics

Message list

Logbook

Device description

FRONT M 700-011

BASE M 700-021

Module Cond Ind 7700

Return Set favorite

0.003 mS/cm

22.7 °C

Device description

FRONT M 700-011

Operating panel M 700

Hardware: 2, Software: 6 .0

Serial number: 0000815

Diagnostics

Select Device description using arrow keys, confirm with enter.

Device description

Please specify the serial number/hardware version when ordering an additional function.

Module FRONT BASE

Return

5

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 6 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW

Activate additional function

Select menu: Parameter setting/System control/Release of options

Note: The TAN for releasing an additional function is only valid for the device with the corresponding serial number (see previous page).

Menu Display Activate additional function

6

Menu selection

7.00 pH

25.6 °C

Select:

Return to meas

[enter]

Lingua

25.6 °C

0.003 mS/cm

Parameter setting

Viewing level

Operator level

Admin. level

(All Data) view

(Operation Data) opl

(All Data) adm

Menu selection

Call up parameter setting.

From the measuring mode:

Press menu key to select menu.

Select parameter setting using arrow keys, confirm with enter.

Parameter setting

Select Administrator level using arrow keys, confirm with enter.

Enter passcode and confirm

(Passcode as delivered: 1989).

Return

16.9 %Air

25.6 °C

Parameter setting (Administrator)

System control

FRONT M 700-011

BASE M 700-021

Module O2 4700

Module pH 2700

Module Cond 7700

Return

0.003 mS/cm

25.6 °C

Release of options (Administrator)

SW 700-001 KI recorder

SW 700-002 Buffer sets

SW 700-003 Adaptive timer

SW 700-004 ServiceScope

SW 700-005 Cal tolerance band

SW 700-006 Current characterist.

Return

Active

Inactive

Inactive

Inactive

Inactive

Inactive

Select System control using arrow keys, confirm with enter.

Then select Release of options using arrow keys, confirm with

enter.

Release of options

Select the additional function to be released.

Set option to “active”. Enter the

TAN at the prompt. (Note: The TAN only applies for the device with the corresponding serial number, see previous page.) The option is available after the TAN has been entered.

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 7 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW

SW 700-001: KI recorder

Select menu: Parameter setting/FRONT M 700/KI recorder

Additional function SW 700-101

The KI recorder follows the course of the process and releases a messages in the event of a deviation. It is called up from the measurement recorder: meas.

Continuous processes

The measured signal is surrounded by a dynamically tracked, processcontrolled tolerance band. Slight fluctuations are tolerated. If the measured value leaves the tolerance range, a message is released (Fig.).

The recording interval can be set from 10 s to 60 h.

Variable Message

Tolerance band around the measured signal generated by a low-pass filter

Measured signal

Fig.: KI recorder, continuous process t

Batch processes

The KI recorder provides an automatic self-learning function. This allows checking whether regularly recurring processes keep the preset tolerances.

Deviations from the normal course of process (plus preset tolerances) generate a message (Fig.). Maximum process duration can be set from 1 to 60 h.

Variable Start End

Preset tolerance band around the measured signal

Measured signal

Fig.: KI recorder, Batch process (Self-learning function) t

7

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 8 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW

SW 700-001: KI recorder

Note: Function check is active during parameter setting.

Menu Display Configure KI recorder

5 pH 9 pH

22 °C 32 °C pH

8,56 pH, 29.4

°C

09.07.02 11:23:33

KI recorder

Info on KI recorder

Recording data

Wait - Function check

Failure right value

Failure left value

Keys :

[enter]

[meas]

[menu]

Cursor to current

To meas. display

Call up menu

Info

°C

Cursor 1 line

Cursor 1/2 page

Close

KI recorder

Called up directly from the measuring mode:

Press meas key.

For function description, press

“Info” softkey

“Info” softkey

Explains icons and capabilities of the

KI recorder

7.00 pH

25.6°C

KI recorder (Administrator)

KI recorder

Channel 1

Channel 2

Off

Continuous process

Batch process

Recording interval

Monitoring

Return

10s

Failure

Configure KI recorder

• Parameter setting FRONT M 700

• Select KI recorder

• Select the variables to be displayed as well as the start and end values

• Set recording interval

(10 s to 60 h) (per pixel)

• Monitoring

(Off/Failure/Maintenance request)

7.00 pH

25.6 °C

Function control matrix (Administrator)

ParSet KI rec.

Fav

– Input OK2

Left softkey

Right softkey

Profibus DO 2 –

Return Connect

Control KI recorder

Example: Configure softkey for

KI recorder start/stop:

• Select parameter setting, then:

• System control

• Function control matrix

• Softkey: “KI rec. Start/Stop”

8

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 9 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_SW

SW 700-001: KI recorder

Menu Display KI recorder: Batch process

7.00 pH

25.6°C

KI recorder (Administrator)

KI recorder

Monitoring

Off

Channel 1 Continuous process

Channel 2

Max. process duration

Batch process

New record Yes No

Failure

Return

Configure KI recorder

• Parameter setting: “Module FRONT”

• Select KI recorder

• Select the variables to be displayed as well as the start, end values, and deviation (%)

• Set process duration

(200 samples are recorded)

• Select “New record Yes”

• Monitoring

(Off/Failure/Maintenance request)

Call up KI recorder

From the measuring mode: Press meas.

Parameters adjustable for batch operation

The gray envelope represents the deviation [%]

End

Variable

Start

9

I pH

°C

I

5

25

15

50 100 150 200

Channel 1

Channel 2

0

Record

50 100 150

KI stop

200

Max. process duration

When the parameters are edited once more, the reference curve is automatically re-adjusted to start, end, and deviaton.

When variable or process duration are changed, the curve will be deleted!

9

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 10 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-001: KI recorder

Display

9

I pH

5

27

°C

I

20

50

0

KI pause

50

100 150

100 150

KI stop

200

200

KI recorder: Batch process

Record reference curve

Start “Record” with left softkey. Reference value and envelope are displayed during the recording process. At the end of recording or after “KI stop” the KI recorder switches to control mode.

9

I pH

5

27

°C

I

50

20

0

KI start

50

100 150 200

In control mode only the envelope is visible.

9

I pH

5

27

°C

I

50

20

0

KI pause

50

100

Cursor

150

KI stop

200

100 150

100 150

KI stop

200

200

Batch control

After a synchronous start the recorder monitors whether the currently measured values remain within the envelope. Every deviation can trigger a message.

KI pause: Stops the recorder.

KI stop: Terminates control.

Pressing “Continue” restarts a stopped recorder.

Note: During “function check” (e.g. during calibration) neither recording nor checking of a batch are interrupted. The last measured value continues to be used.

10

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 11 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-001: KI recorder

Display KI recorder: Batch process

9

I pH

5

50

Evaluation channel 1:

End 9.00 pH

Start

Deviation

Process duration

5.00 pH

10 %

10 h

KI pause

100 150 200

Max

Min

Max. deviation

Messages

KI stop

8.35 pH

5.80 pH

7 %

0x

°C

I

9

I pH

5

27

20

50 100

0 50

Control via OK2

100

150

150

200

200

7.00 pH

25.6 °C

Contact K3

Usage

Contact type

NAMUR maintenance!

NAMUR function check

Limit

Rinse contact

KI rec active

Return

Evaluation

Within the KI recorder function you can evaluate a reference curve directly on the device.

To do so, use the arrow keys (Up/Down). The reference curves and the most important parameters are displayed one after the other.

External control

The Start/Stop function can be remote-controlled via the OK2 input (BASE module) or via the PA 700 (X) PROFIBUS module.

(To set the control element, call up:

Parameter setting / System control / Function control matrix).

The softkeys will then be deactivated.

The K1, K2, and K3 contacts of the BASE module can be assigned to the “KI rec active” signal.

• Save / load reference curves

Up to 5 batch reference curves can be stored on and retrieved from a SmartMedia card:

Parameter setting / System control / KI batch recording.

The reference curve, envelope, time response, and KI settings are stored.

• Archiving of completed batch processes

Complete batch results can be archived on the SmartMedia card. All reference values and current batch data are stored in the archive.

11

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 12 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

12

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 13 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-002: Buffer set to be entered (pH)

Select menu: Parameter setting/System control/Buffer table

Individual buffer set (with 3 buffer solutions) for pH measurement

Buffer table

You can enter an individual buffer set (with 3 buffer solutions).

To do so, you enter the nominal buffer values for the correct temperature

(range 0 ... 95 °C, 5°C steps). Then this buffer set is available in addition to the permanently set standard buffer solutions in the “Calimatic buffer” menu (select “Table”).

Menu Display

7.20 pH

25.6 °C

System control (Administrator)

Calculation Blocks

Point of measurement

Release of options

Logbook

Buffer table

Concentration chart

Return

Buffer table (Administrator)

Buffer 1

Buffer 2

Buffer 3

7.10 pH

25.6 °C

Buffer table: Enter values

Enter buffer set

• Call up parameter setting

• System control

• Select “Buffer table”

• Select buffer to be entered

Return

7.10 pH

25.6 °C

Buffer 1 (Administrator)

Nominal buffer value pH value at 00 °C pH value at 05 °C pH value at 10 °C pH value at 15 °C pH value at 20 °C

Return

+04.00 pH

+04.00 pH

+04.00 pH

+04.00 pH

+04.00 pH

+04.00 pH

• Enter nominal buffer value and all other values for the correct temperature (right/left arrow keys to select position, up/down arrow keys to edit number, confirm with

enter.)

The special buffer set is selected as follows:

Parameter setting/Module pH/Cal preset values/Calimatic buffer/Table.

13

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 14 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-004: ServiceScope (pH)

Select menu: Diagnostics/Module pH 2700/ServiceScope

ServiceScope

ServiceScope monitors the pH input signal and checks whether it remains in the control range. An error message is generated if the input signal exceeds the failure limit.

Menu Display

ServiceScope

Fail

Noise level

Return

ServiceScope

7.00 pH

0.2 °C t

ServiceScope

(Diagnostics/Module pH 2700)

Displays the noise levels over the time. Allows distinction of individual disturbances, periodic and broadband disturbances, which is helpful for troubleshooting. An error message is generated if the noise level exceeds the failure limit.

14

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 15 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-005: Tolerance band recorder (pH)

Select menu: Diagnostics/Module pH/Tolerance band recorder

Tolerance band recorder

(Adjustment of tolerance band: see following page)

Tolerance band calibration prevents that slight calibration scatter of zero and slope immediately leads to a readjustment of the calibration data and thus to a shifting of the measured value. If the zero and slope values determined during calibration remain within their tolerance bands, the new data are not stored. If one of the calibration values lies outside the tolerance band, the new data are taken over.

Drift due to aging or calibration scatter can be identified at a glance, thus allowing to draw conclusions as to electrode life and the required calibration interval.

In the Calibration record you see whether the data have been stored or whether the old data can still be tolerated.

Menu Display

Tolerance band recorder

Cal record

6.00

T-band recorder

ZERO 8.00

50.7

Return

7.00 pH

22.3 °C

Slope 61.7

Tolerance band recorder

(Diagnostics/Module pH 2700)

Records the tolerance ranges for zero and slope over the time. New calibration values are only accepted if the tolerance limits have been exceeded.

Display can be graphical or as a listing.

7.00 pH

22.3 °C

Cal record

Last calibration

Cal tolerance band

Cal mode

Zero point

Slope

Isothermal potential

Return

01.05. 2003 11:24

Old el data

7.00 pH

58.0 mV/pH

0 mV

Cal record

(Diagnostics/Module pH 2700)

In the Cal record you can see whether the data have been stored

(New el data) or whether the old calibration data can still be tolerated

(Old el data).

15

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 16 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-005: Tolerance band recorder (pH)

Adjustment of tolerance band. Select menu:

Parameter setting/Module pH 2700/Cal preset values/Cal tolerance band

Menu Display

Menu selection

Select:

Return to meas

7.0 pH

17.1 °C

[enter]

Lingua

Parameter setting

Menu selection

From the measuring mode:

Press menu key to select menu.

Select parameter setting using arrow keys, confirm with enter.

Passcode as delivered: 1989

7.00 pH

17.1°C

Parameter setting (Administrator)

System control

FRONT M 700-011

BASE M 700-021

Module Cond 7700

Module PID 700

Module pH 2700

Return

7.00 pH

17.1 °C

Module pH 2700 (Administrator)

Input filter

Sensor data

Cal preset values

TC process medium

ORP/rH value

Delta function

Return Block

7.00 pH

17.1 °C

Cal tolerance band (Administrator)

Calibration data taken over when tolerance band exceeded

Tolerance band check

Tolerance Zero

Tolerance Slope

On Off

+02.00 pH

002.0 mV/pH

Return Block

Parameter setting

Select “Module pH 2700”, confirm with enter.

pH module

Select “Cal preset values” using arrow keys.

Confirm with enter.

Then select “Cal tolerance band”.

Cal tolerance band

Enter tolerance ranges for zero and slope

(enter numbers using arrow keys, confirm with enter.)

16

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 17 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-006: Current characteristic definable

Select menu: Parameter setting/BASE M 700 (Out)/Output current/Curve

Menu Display

7.03 pH

25.6 °C

Output current I1 (Administrator)

Variable

Curve

Output

Start

End

Output filter

Abort pH

Linear

Trilinear

Function

Chart

0000 s

OK

7.03 pH

25.6 °C

Output current I1 (Administrator)

Variable

Curve pH

Chart

Chart

Output 4 ... 20 mA

Output filter 0000 s

Behavior during messages

Abort OK

Specify current characteristic

Chart characteristic

Assign output current to measured value in 1 mA steps.

Select menu: Parameter setting/

BASE M 700 (Out)/Output current

First select process variable!

Then curve: Chart

The “Chart” menu is displayed.

7.03 pH

25.6 °C

Chart (Administrator)

Value at 00 mA:

Value at 01 mA:

Value at 02 mA:

Value at 03 mA:

Value at 04 mA:

Value at 05 mA:

Abort

+ 00.00 pH

+ 00.00 pH

+ 00.00 pH

+ 00.00 pH

+ 00.00 pH

+ 00.00 pH

OK

The entered values must be continuously rising or falling.

On the following page, you find a listing to write down your adjustments.

17

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 18 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

Current curve values - Original for copy

Parameter setting/BASE M 700 (Out)/Output current/

Curve: Chart, individual settings

Value at measured variable

00 mA ............... _________

01 mA ............... _________

02 mA ............... _________

03 mA ............... _________

04 mA ............... _________

05 mA ............... _________

06 mA ............... _________

07 mA ............... _________

08 mA ............... _________

09 mA ............... _________

10 mA ............... _________

11 mA .............. _________

12 mA .............. _________

13 mA ............... _________

14 mA ............... _________

15 mA ............... _________

16 mA ............... _________

17 mA ............... _________

18 mA ............... _________

19 mA ............... _________

20 mA ............... _________

18

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 19 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-008: TC ultrapure water (Cond)

Select menu:

Parameter setting/Module Cond 7700/TC process medium/Ultrapure water

Menu Display TC process medium

0.20 mS/cm

22.3 °C

TC process medium (Administrator)

EN 27888: Natural waters (0 ... 35°C)

Ultrapure water: traces of impurity

TC correction

Reference temp

Impurity

Ultrapure water

Return OK

NaOH

NaCl

HCl

NH3

Ultrapure water with traces of impurity

• Parameter setting

• Module Cond 7700

• TC process medium

• TC correction: Ultrapure water

Select type of impurity:

• NaOH

Alkaline ultrapure water

• NaCl

Neutral ultrapure water, for conductivity measurement in water processing behind gravel bed filter

• HCl

Acidic ultrapure water, for conductivity measurement behind cation filter

• NH3

Ammoniacal ultrapure water

19

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 20 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-009: Concentration determination

Select menu: Parameter setting/System control/Module Cond/Concentration

Default settings and selection range

Concentration determination

The substance concentration in percent by weight (% by wt) is determined for H

2

SO

4

, HNO

3

, HCl, NaOH, NaCl from the measured conductivity and temperature values.

Conditions for concentration determination

Below you find tables with the concentration ranges of the preset substances.

These tables show the conductivity of the substances versus substance concentration and process temperature:

• For calculation of concentration, the medium to be measured must be a purely binary mixture (e.g. water-hydrochloric acid). Presence of other dissolved substances (e.g. salts) leads to incorrect concentration values.

• In the region of small slopes (e.g. at the range limits) small conductivity changes can correspond to great concentration changes. This may lead to an unsteady display of the concentration value.

• As the concentration value is calculated from the measured conductivity and temperature values, accurate temperature measurement is very important. Therefore, you should make sure that sensor and process medium are in thermal equilibrium.

Messages

You can define limits for warning and failure messages for the concentration value:

Select menu: Parameter setting/System control/Module Cond/Messages

20

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 21 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

Parameter setting: Concentration curves

(Additional function SW 700-009) Default settings and selection range

Note: Function check active

Parameter Default Selection / Range

Concentration

• Medium (“Yes” selected)

No

H2SO4 (0-30%)

USP function

• Reduced limit

• Monitoring

Off

100 %

Off

Yes, No

H2SO4 (0-30 %), H2SO4 (32-84 %), H2SO4 (92-99 %),

HNO3 (0-30 %), HNO3 (35-96 %),

HCl (0-18 %), HCl (22-39 %)

NaOH (0-14 %), NaOH (18-50 %),

NaCl (0-26 %),

Chart (additional function SW 700-009)

Off, On

10 % ... 100%

Off, failure, maintenance request

Sulphuric acid H

2

SO

4 c[% by wt]

21

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 22 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

Nitric acid HNO

3 c[% by wt]

Hydrochloric acid HCl

22

c[% by wt]

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 23 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

Sodium hydroxide solution NaOH c[% by wt]

Table salt solution NaCl c[% by wt]

23

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 24 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-009: Concentration chart (Cond)

Select menu: Parameter setting/System control/Concentration chart

Specifying a concentration solution for conductivity measurement

Concentration chart

To specify the customer-specific solution, 5 concentration values A-E are entered in a matrix together with 5 temperature values 1-5. To do so, first enter the 5 temperature values, then enter the respective conductivity values for each concentration A-E.

These solutions will then be available in addition to the permanently stored standard solutions (select “Chart”).

Menu Display Enter concentration chart

0.020 mS/cm

25.6 °C

System control (Administrator)

Calculation Blocks

Point of measurement

Release of options

Logbook

Concentration chart

Factory setting

Return

Enter values

• Call up parameter setting

• System control

• Select “Concentration chart”

0.020 mS/cm

25.6 °C

Concentration table (Administrator)

Temp 1

Temp 2

Temp 3

Temp 4

Temp 5

+000.0 °C

+005.0 °C

+010.0 °C

+015.0 °C

+020.0 °C

Abort OK

0.020 mS/cm

25.6 °C

Concentration table (Administrator)

Concentration A:

1. Cond at +000.0 °C

2. Cond at +005.0 °C

3. Cond at +010.0 °C

4. Cond at +015.0 °C

5. Cond at +020.0 °C

Abort

05.00 % by wt

0.000 µS/cm

0.000 µS/cm

0.000 µS/cm

0.000 µS/cm

0.000 µS/cm

OK

Enter 5 temperature values

(right/left arrow keys to select position, up/down arrow keys to edit number, confirm with enter.)

Enter values for concentrations A-E for the respective temperatures. The table values must be continuous.

Maxima/minima are not permitted.

Incorrect entries are marked with

The concentration chart is selected as follows:

Parameter setting/Module Cond/Cal preset values/Automatic/Chart.

xx .

24

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 25 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

The chart is built up as 5 x 5 matrix:

Temp 1

Conc. A Conc. B Conc. C Conc. D Conc. E

A1 B1 C1 D1 E1

Temp 2

Temp 3

A2

A3

B2

B3

C2

C3

D2

D3

E2

E3

Temp 4

Temp 5

A4

A5

B4

B5

C4

C5

D4

D5

E4

E5

Conditions for the chart:

• The temperature values must be rising (Temp 1 is the lowest, Temp 5 the highest temperature).

• The concentration values must be rising (Conc. A is the lowest, Conc. E the highest concentration).

• The chart values A1 ... E1, A2 ... E2 etc. must all be rising within the chart or all falling. Points of inflection are not allowed!

The M 700 automatically checks the chart values. Incorrect entries are marked.

25

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 26 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-011: High CO2 compensation

Application-specific additional function for breweries

This additional function simplifies parameter setting since all steps not required for measurement of beer are omitted. It simultaneously acts on all installed O

2 module (module software version 2.2 and higher).

Function principle:

During the filling process it must be ensured that as little oxygen as possible is dissolved in the beer to extent its shelf life.

During oxygen trace measurement the sensor is operated with a very low polarization voltage (-500 mV). This results in low cross-sensitivity to CO

2

.

For a calibration in air, this polarization voltage is too low.

It must be set to -675 mV and afterwards be reduced again to -500 mV for measuring in the trace range.

A sufficiently long waiting time must be observed until the sensor has stabilized.

Opening and closing of valves causes pressure variations in the beer pipes which momentarily falsify the O

2 signal. Therefore the input signal must be attenuated correspondingly to suppress transient interferences.

These processes are automated by the additional function, i.e. all parameters required for the respective program step are set automatically. When the additional function has been activated, the parameter setting steps are displayed as follows:

26

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 27 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

Parameter Default Selection / Range

Input filter

• Pulse suppression

Sensor data

• Measure in

• Sensor type

• Temperature probe **

• Sensor **

• Reference electrode **

• Membrane correction

• Polarization voltage

• Sensocheck

Cal preset values

• Cal saturation

• Cal concentration

• Calibration timer

Pressure correction

• Pressure during meas

• Pressure during cal

Salt correction

• Input

Weak

Liquids

A Standard

NTC 22 k Ω

Without guard

Off

1.00

0500 / 0675 mV

Off

A Standard, B Trace sensor (with guard) * ,

C Trace sensor (without guard), Others

NTC 30 kOhm, NTC 22 kOhm

With guard, Without guard

Off, On

0.50 ... 2.00

For sensor type A, B, C: fixed

Off, Failure, Maintenance request

%AIR mg/l

0000 h

Auto

Auto

Off, Weak, Moderate, Strong, On (000 s ... 999 s)

%Air

µg/l, mg/l, ppb, ppm xxxx h (entry)

Auto, Manual (default value 1013 mbars)

Auto, Manual (default value 1013 mbars)

Salinity

Salinity, Chlorinity, Conductivity

(00.00 g/kg or 0.000 µS/cm, depending on selection)

* Sensocheck not possible for trace sensor with guard, therefore disabled

** Can only be set with “Sensor type Others” selected

27

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 28 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-012:

ISFET for ISM module pH 2700i(X)

pH measurement with InPro 3300 ISFET sensor pH 2700i(X) jumper

Note:

Each time a new sensor is connected, an ISFET zero point adjustment must be performed.

After that, you should perform one of the following calibration methods:

• Calimatic: automatic calibration

• Manual: entry of buffer values

• Data entry: premeasured electrodes

28

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 29 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

Parameter settings: New

Modul pH 2700i (Administrator)

7.00 pH

20.1 °C

Input filter

Sensor data

Cal preset values

TC process medium

ORP/rH value

Delta function

Return Block

7.00 pH

20.1 °C

Sensor data (Administrator)

Sensor type ISFET

Temperature detection

Sensoface

Sensor monitoring details

(Pt100)

On Off

Abort OK

7.00 pH

20.1 °C

Sensor monitoring details (Administrator)

Slope

Zero

Sensocheck ref el

Sensocheck glass el

Response time

ISFET leak current

Return

(Auto)

(Auto)

(Auto)

(Auto)

(Auto)

(Auto)

Parameter setting/ module pH 2700i / Sensor data

Sensor data are preset depending on the sensor type.

Gray display lines cannot be edited.

Sensor data

Option SW 700-012 allows the setting of “ISFET” as Sensor type.

Sensor monitoring details

Monitoring of ISFET leak current is available.

Sensor monitor pH input

ORP input

RTD

Temperature

Impedance ref (25°C)

ISFET characteristic

Return

7.00 pH

22.3 °C

43 mV

-109 mV

1.100 k Ω

25.6 °C

322.8 k

Ω

0,81 mA/V

7.00pH

24.1°C

Sensor network diagramm PH

1 - Slope

2 - Zero point

3 - Ref mpedance

4 - ISFET Leak current

5 - Response time

6 - Calibration timer

7 - Sensor wear

Return

Diagnostics: New

Sensor monitor

ISFET characteristic.

Sensor network diagramm pH

New: ISFET leak current.

29

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 30 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

Inserting the SmartMedia card

To release an additional function via TAN, see Pg 6, Release of options

Inserting and replacing the SmartMedia card

The SmartMedia card may be inserted or replaced with the power supply switched on. Protect against electrostatic discharge! When closing the device, make sure that the sealing is properly seated and clean.

Warning!

Do not touch the terminal compartment, there may be dangerous contact voltages!

1. To open the analyzer

• Loosen the 4 front screws

• Open the M 700 FRONT at its right side

(pivot hinge inside at the left)

• The slit for inserting the

SmartMedia card is located at the inner side of the M 700 FRONT module

2. To insert the SmartMedia card

• Remove SmartMedia card from its package without touching the contact surface

• Insert card in the slit at the inner side of the

M 700 FRONT module

Inserting the

SmartMedia card:

The label must be facing you.

30

3. To remove the SmartMedia card

• Please call up the Maintenance menu to avoid data loss.

• “Close memory card” to terminate software access to the SmartMedia card. Now the card can be taken out.

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 31 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SmartMedia card: Usage

Use as memory card in combination with additional functions.

Additional functions must be ordered separately (release via TAN).

A SmartMedia card is provided with the following additional functions:

• Software update (SW 700-106, see Pg 43)

• Additional function SW 700-1xx (description Pg 34 et seq.)

Using commercially available SmartMedia cards

Commercially available SmartMedia cards can be used as additional memory card.

The following types of cards are supported: 8 MB, 16 MB, 32 MB, 64 MB,

128 MB. Externally produced files, such as from a digital camera, are tolerated. Long file names can be read. The M 700 generates file names in the

8.3 format (8 characters file name, 3 characters program-specific file name extension).

Formatting a commercial SmartMedia card

Some devices (e.g. digital cameras, scanners) cause a formatting of the

SmartMedia card which does not correspond to the SSFDC specification or the SmartMedia Interface Library (SMIL). The manufacturer therefore recommends to format a commercial SmartMedia card as M 700 memory card prior to first use.

Menu Display Format SmartMedia card

0.003 mS/cm

20.4 °C

Memory card (Administrator)

Record logbook On

Record recorder

Decimal separator

Card full

On

Point

Record stop

Off

Off

Comma

Data saving

Format card

Return

Formatting

• Insert SmartMedia card

• Open menu selection

• Parameter setting, Admin. level

• Enter passcode

• System control: Memory card

(The “Memory card” function is only available with the SmartMedia

Card inserted!)

• Format card

31

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 32 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

Loading /saving configuration

System control/Memory card/Copy configuration.

Additional functions must be ordered separately (release via TAN).

Saving / loading the complete device configuration

System control/Memory card/Copy configuration.

“Save“ configuration means that the complete device configuration (except the passcodes) is written on the memory card.

“Load“ configuration means that the complete device configuration is read from the memory card and pogrammmed in the M 700.

BACKUP file generated on SmartMedia card: \BACKUP\BACKUP01.PAR

Copy configuration from one M 700 to further M 700 units

(Copying parameter sets for similarly equipped M 700)

Prerequisite:

The devices must have the same hardware.The module equipment must be absolutely identical – the same modules in the same slots!

Options:

All required options must be enabled on the “master device”, this is not required for the“slave devices”.

It is the parameters of the options that are transferred, not the option itself.

When an option will be enabled on a “slave device” at a later point in time, the parameters of this option will be preset corresponding to the “master device”.

1 Write device configuration of configured device on SmartMedia card:

Parameter setting/System control/Copy configuration/Save.

2 Call up Maintenance menu. Select “Close memory card”.

3 Remove SmartMedia card.

Now you can transfer the device configuration to further identically equipped devices.

4 To do so, insert the SmartMedia card containing the configuration in the next device to be configured. Select

Parameter setting/System control/Copy configuration/Load.

5 Call up Maintenance menu. Select “Close memory card”.

6 Remove SmartMedia card.

32

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 33 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SmartMedia card: Format card

Parameter setting/System control/Format card

Menu Display

Format card

(Generate memory card)

0.003 mS/cm

20.4 °C

Software update (Administrator)

Updating will change the device characteristics. Perform verification as appropriate!

Update firmware

Save firmware

Format card

Return

0.003 mS/cm

20.4 °C

Format card (Administrator)

Attention: All data on the memory card are deleted!

Return Format

Format card

• Insert SmartMedia card (Pg 30)

• Open menu selection

• Parameter setting, Admin. level

• Enter passcode

• System control: Format card

Caution!

Formatting an update card generates a memory card.

Refer to “SmartMedia card”, Pg 34.

This process is irreversible! Double warning messages proect against faulty operation.

Note:

The SmartMedia card must always be formatted in the M 700.

Correct functioning of a SmartMedia card with different formatting cannot be guaranteed.

File structure of a memory card (example):

Folder

BACKUP

RECORDER

Typ. file name

BACKUP01.PAR

LOGBOOK L_YYMM00.TXT

PARASET SET 1

R_YYMMDD.TXT

Remark

Parameter set (Backup)

Logbook file,

YY=year, MM=month

Parameter set

Recorder entry,

YY=year, MM=month, DD=day

33

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 34 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SmartMedia card: Memory card

Parameter setting/System control/Memory card

Menu Display

34

Using the memory card

0.003 mS/cm

20.4 °C

Software update (Administrator)

Updating will change the device characteristics. Perform verification as appropriate!

Update firmware

Save firmware

Format card

Return

System control (Administrator)

Memory card

Copy configuration

Parameter sets

Function control matrix

Time/date

Point of measurement

0.003 mS/cm

25.6 °C

Return

0.003 mS/cm

25.6 °C

Memory card (Administrator)

Record logbook On

Record recorder On

Decimal separator

Card full

Point

Record stop

Format card

Off

Off

Comma

Return

Using the memory card

• Insert SmartMedia card (Pg 30)

• Open menu selection

• Parameter setting, Admin. level

• Enter passcode.

• System control: Memory card

With SmartMedia card inserted, the display shown on the left appears

(The “Memory card” line is displayed only if there really is a memory card in the slot).

• Select “Memory card”, confirm with enter.

The menu is self-explanatory.

Behavior when the memory card is full:

Continuous recording (as with a flight recorder) or Stop (card replacement).

0.003 mS/cm

25.6 °C

Copy configuration (Administrator)

Identical module equipment required for transfer.

Configuration

Return

Save

Load

Save

Copy configuration

• Save: saves all data on the memory card

• Load: Overwrites all device data with the data from the memory card

Caution!

“Close” memory card before removing it (Maintenance menu)

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 35 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-102: Loadable parameter sets

Parameter setting/System control/Parameter sets

Menu Display

Load parameter set from

SmartMedia card

Parameter sets (Administrator)

0.003 mS/cm

25.1 °C

Select with softkey

Settings in menu

Softkey usage

Parameter sets

Save parameter set A

Load parameter set A

A,B (internal)

1,2 (card)

1,3 ,3 (card)

1,2,3,4 (card)

Abort

1,2,3,4,5 (card)

OK

Parameter sets (Administrator)

0.003 mS/cm

25.0 °C

Select with softkey

Settings in menu

Softkey usage

1,2,3,4 (card) Parameter sets

Save parameter set A

Load parameter set A

Return

Loading a parameter set from

SmartMedia card

• Call up parameter setting

• System control

• Call up “Parameter sets” (Fig)

The M 700 provides 2 complete parameter sets (A, B).

5 parameter sets can be stored on the SmartMedia card. One of those can be saved as parameter set A to the M 700:

SmartMedia card 0.003 mS/cm

24.9 °C

Load parameter set A (Administrator)

NOTE: Old parameter set will be overwritten!

B (internal)

1 (card)

Load from 2 (card)

3 (card)

Abort

Set 1 Set 2 Set 3

Set A

Set B

(internal only)

0.003

25.0

mS/cm

°C

Time 11:49 Favorites menu

• Select parameter set to be loaded

Activated parameter set is displayed in measuring mode.

Note: Remote switching between A and B is possible via the OK2 input.

35

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 36 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-102: Loadable parameter sets

Parameter setting/System control/Parameter sets

Menu Display

Save parameter set on

SmartMedia card

Parameter sets (Administrator)

0.003 mS/cm

25.1 °C

Select with softkey

Settings in menu

Softkey usage

Parameter sets

Save parameter set A

Load parameter set A

Abort

A,B (internal)

1,2 (card)

1,2,3 (card)

1,2,3,4 (card)

1,2,3,4,5 (card)

Parameter sets (Administrator)

0.002 mS/cm

24.2 °C

Select with softkey

Settings in menu

Softkey usage

1,2,3,4 (card) Parameter sets

Save parameter set A

Load parameter set A

Return

To save a parameter set on

SmartMedia card

• Call up parameter setting

• System control

• Call up “Parameter sets” (Fig)

The M 700 provides 2 complete parameter sets (A, B).

Up to 5 parameter sets can be loaded to the SmartMedia card.

To do so, a parameter set

(1, 2, 3, 4, or 5) of the SmartMedia card is overwritten by parameter set

A of the M 700.

0.002 mS/cm

24.2 °C

Save parameter set A (Administrator)

NOTE: Old parameter set will be overwritten!

B (internal)

1 (card)

Save to 2 (card)

3 (card)

Abort

4 (card)

5 (card)

• Select parameter set on SmartMedia card

Set A

Set B

(internal only)

Set 1 Set 2 Set 3

SmartMedia card

36

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 37 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-103: Measurement recorder

Select menu: Parameter setting/FRONT M 700/Measurement display

The last 200 data records are stored in the M 700 and are represented graphically – like on a recorder. All further data can be stored on the

SmartMedia card and be evaluated by a computer.

2 process variables are represented. You can configure:

• The process variable to be displayed

• Start and end value for the variable to be recorded

• Time base (recording interval, selectable from 10 s to 10 h)

In addition, the time axis can be stretched by factor 10 using the “Zoom function”. In the event of quick changes, the zoom function is automatically activated. It always starts several pixels before the event. This allows detailed investigation of signal instabilities.

With “Min/Max displays” activated, the respectively measured peak values

(gray) are shown in addition to the measured value (average).

0.002 mS/cm 0.004 mS/cm 20°C 25°C

Graphic range

Start and end value of process variable

S/cm

07/24/02 11:32, Status: Measurement

0.003mS/cm, 23.9°C

1 2

Info

°C

10x

Events (NAMUR) (see “Info”)

Maintenance request

Function check

Failure

Min/Max display

The minimum and maximum values within the recording interval are shown as a gray line next to the (averaged) measured value.

Fig. 1: Min/Max off

Fig. 2: Min/Max display

37

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 38 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-103: Measurement recorder

Select menu: Parameter setting/FRONT M 700/Measurement display

Additional function SW 700-103: Measurement recorder

The measurement recorder saves all entries in a file. The last 200 entries can be displayed on the M 700. A new file is generated for each day. The date is encoded in the file name.

Example for a file generated on SmartMedia card:

\RECORDER\R_YYMMDD.TXT

Recorder data of YYMMDD

(YY = year, MM = month, DD = day)

The data is recorded as ASCII file with the extension .TXT.

The individual columns are separated by tabs. This makes the file readable with word processing or spreadsheet programs (e.g. Microsoft Excel). Each time the memory card is inserted in the slot, a “Device Info” consisting of

Model number, BASE serial number, and tag number. Thus, a memory card can also be used to collect the measurement recorder data of several devices.

Example:

M 700 - Measurement recorder

Time stamp CH Left

<< M 700 - Serial 002004 [DSE KL_007] >>

21.10.02 00:09:15

21.10.02 00:19:15

21.10.02 00:29:15

21.10.02 00:39:15

21.10.02 00:49:15

21.10.02 00:59:15

1015

1015

1015

1015

1015

1015 mbar mbar mbar mbar mbar mbar

2.8

2.9

2.8

2.9

2.9

3.0

CH Right

°C

°C

°C

°C

°C

°C

M-req Fct.

0

0

0

0

0

0

0

0

0

0

H

0

Fail

0

0

0

0

H

0

Slow

0

0

0

0

0

0

38

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 39 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-103: Measurement recorder

Select menu: Parameter setting/FRONT M 700/Measurement display

Meaning of the entries in the recorder file:

Time stamp Time stamp of recorder entry

CH Left Left channel of recorder with measured value and unit

CH Right

M-req.

Fct.

Fail

Slow

Right channel of recorder with measured value and unit

NAMUR signal* “Maintenance request”

NAMUR signal* “Function check”

NAMUR signal* “Failure”

Zoom function active (Slow motion)

* Definition:

NAMUR = German committee for measurement and control standards in the chemical industry

39

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 40 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-104 : Extended logbook

Parameter setting/System control/Logbook

Additional function SW 700-104: Extended logbook

The extended logbook saves all entries in a file. The last 50 entries can be displayed on the M 700. A new file is generated for each month. The date is encoded in the file name.

Example for a file generated on SmartMedia card:

\LOGBOOK\L_YYMM00.TXT

Recorder data of YYMM

(YY = year, MM = month)

The data is recorded as ASCII file with the extension .TXT.

The individual columns are separated by tabs. This makes the file readable with word processing or spreadsheet programs (e.g. Microsoft Excel). Each time the memory card is inserted in the slot, a “Device Info” consisting of

Model number, BASE serial number, and tag number is written. Thus, a memory card can also be used to collect the logbook data of several devices.

Example:

M 700 - Logbook

Time Stamp Status Message

<< M 700 - Serial 0002004 [DSE KL_001] >>

21.11.02

19:08:43

22.11.02

06:02:01

22.11.02

06:09:27

22.11.02

06:09:36

23.11.02

16:45:07 (x)

Power supply Off

Power supply On

Diagnostics active

Measurement active

Fail current I2 > 20 mA

23.11.02

18:43:11

23.11.02

18:47:38

23.11.02

18:47:38 ( )

:

:

Parameter setting active

Measurement active

Fail current I2 > 20 mA

Time stamp:

Status

Message

40

Time stamp of logbook entry

(x) - Message activated

( ) - Message deactivated

Message text (in selected operator language)

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 41 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-106: Software update

For a software update (additional function SW700-106), the manufacturer supplies a specially formatted SmartMedia card. The M 700 replaces its own firmware (operating program) by the new version (“Update”).

Caution!

During a software update the M 700 is not operable!

After a software update you should check the configuration.

UP

This icon indicates that a SmartMedia card is inserted in the slot.

The card allows storing of current device software on the card as well as loading of new software into the M 700.

1. Save the firmware currently installed in your M 700 (Pg 42).

2. Load the software update as described on Pg 43.

Note:

A memory card can be generated by formatting an update card (irreversible!).

See Page 31.

41

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 42 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SmartMedia card: Save firmware

Parameter setting/System control/Software update/Save firmware

Menu Display

Save firmware on software update card

UP

0.003 mS/cm

20.4 °C

Software update (Administrator)

Updating will change the device characteristics. Perform verification as appropriate!

Update firmware

Save firmware

Format card

Return

UP

0.003 mS/cm

20.4 °C

Save firmware (Administrator)

UP

Select a free memory slot or overwrite a stored firmware version

Save firmware

• Insert SmartMedia card

• Open menu selection

• Parameter setting, Admin. level

• Enter passcode

• System control: Software update

Select a free memory slot on the card:

Slot 2: SZ34Z10000/0-V1.4

Return Save

UP

0.003 mS/cm

20.4 °C

Save firmware (Administrator)

UP

• Select slot with key

• Select free slot with arrow key.

Slot

5: <empty>

6: <empty>

7: <empty>

Return OK

UP

0.003 mS/cm

20.4 °C

Save firmware (Administrator)

Select a free memory slot

UP or overwrite a stored firmware version

• Confirm slot with “OK”.

Slot 3: <empty>

Return

UP Save successful

Save

Start with “Save” softkey.

Confirm finish message

(with “OK” or enter).

Remove the SmartMedia card.

Close the front door.

42

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 43 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW 700-106: Software update

Parameter setting/System control/Software update/Load firmware

Menu Display

Software update

(“Load firmware”)

0.003 mS/cm

25.6 °C

Release of options (Administrator)

102 Parameter sets

103 Measurement recorder

104 Logbook

106 Software update

107 AuditTrail

Inactive

Inactive

Inactive

Active

Inactive

Software update

• Insert SmartMedia card (Pg 30)

• Open menu selection

• Parameter setting, Admin. level

• Enter passcode

• Select System control

Return

0.003 mS/cm

20.4 °C

Software update (Administrator)

Updating will change the device characteristics. Perform verification as appropriate!

Update firmware

Save firmware

Format card

Return

UP

0.003 mS/cm

20.4 °C

Update firmware (Administrator)

UP

Select the memory slot for

( the chosen firmware indicates the existing version)

Slot 1: SZ34Z10000/0-V1.0

Abort OK

UP

0.003 mS/cm

20.4 °C

Update firmware (Administrator)

UP

Select the memory slot for

( the chosen firmware indicates the existing version)

1: Select Release of options

(Software update SW 700-106)

Set option to “active”. Enter the

TAN at the prompt. The option is available after the TAN has been entered.

2: Select Software update

Check whether your unit really requires a software update!

To read the current software version, select:

• Diagnostics

• Device description

• FRONT M 700

Perform update:

• Parameter setting

• System control

• Software update

• Select slot

• Confirm slot with “OK”.

• Press “Load firmware” softkey to start software update.

Slot 1: SZ34Z10000/0-V1.0

Return Load firmware

43

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 44 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 45 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

AuditTrail and signature

Electronic data recording and protection according to FDA 21 CFR Part 11

Additional function SW 700-107

Provided by: TAN and AuditTrail card

Contents

AuditTrail....................................................................................................46

Signature....................................................................................................47

Activating the AuditTrail .............................................................................48

Signature entry...........................................................................................49

AuditTrail: Parameter setting.......................................................................50

AuditTrail: User management .....................................................................51

AuditTrail: Recording ..................................................................................52

AuditTrail log ..............................................................................................53

45

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 46 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

AuditTrail

Function description

Electronic data recording and protection according to

FDA 21 CFR Part 11

Pharmaceutical applications not only require a hygienic design, but also the protection and complete recording of data (FDA regulation: 21 CFR Part 11).

The M 700 with stainless steel enclosure and SmartMedia card meets these demands. The FDA regulation basically consists of two parts: “Audit Trail” and “Electronic Signature”. Both parts interact directly: For every change of parameters, the device records what has been changed and who has changed it.

Functions like calibration (adjustment) often may only be performed after written approval! This approval is given by an electronic signature.

AuditTrail card

The AuditTrail card is a specially prepared SmartMedia card. The M 700 records the AuditTrail log on an “AuditTrail card” only. Readability from PC is not affected.

Note:

Only specially coded SmartMedia cards supplied from Mettler Toledo can be used as AuditTrail cards.

As delivered, the AuditTrail card includes a PC program which makes the recorded AuditTrail data readable. When log coding has been activated, it also allows decoding of the data. The program supports the data export to other programs, e.g. Microsoft Excel.

46

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 47 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

Signature

Function description

Signature

A signature consists of two entries, the user name and the corresponding passcode. For each change at the device (calibration/adjustment, parameter setting, maintenance, if required also diagnostics), the M 700 records in the

“AuditTrail Log” and in addition in the logbook who has performed the change. For that purpose, access authorizations are introduced.

Access authorization

Every access authorization consists of a signature:

• User name and

• Passcode.

The administrator and the user enter the access data in the user management menu. The administrator enters the user name (short name) and the user adds his/her personal passcode.

Each time he/she calls up a protected menu, the user must enter his/her name and passcode. The M 700 only gives access to the function if the access codes are valid. The recordings (AuditTrail log, logbook) on the

AuditTrail card include the user names so that all actions can be clearly traced back.

Coded storage

The entries in the AuditTrail log can be coded using an algorithm and protected with checksums so that a special evaluation program can prove the authenticity of the data records on a PC.

47

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 48 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

Activating the AuditTrail

Select menu: Parameter setting/System control/Release of options

Note: The TAN for releasing an additional function is only valid for the device with the corresponding serial number

Menu Display Activate additional function

48

Menu selection

7.00 pH

25.6 °C

Select:

Return to meas

[enter]

Lingua

25.6 °C

0.003 mS/cm

Parameter setting

Viewing level

Operator level

Admin. level

(All Data) view

(Operation Data) opl

(All Data) adm

Menu selection

Call up parameter setting.

From the measuring mode:

Press menu key to select menu.

Select parameter setting using arrow keys, confirm with enter.

Parameter setting

Select Administrator level using arrow keys, confirm with enter.

Enter passcode and confirm

(Passcode as delivered: 1989).

Return

16.9 %Air

25.6 °C

Parameter setting (Administrator)

System control

FRONT M 700-011

BASE M 700-021

Module O2 4700

Module PH 2700

Module Cond Ind 7700

Return

0.003 mS/cm

25.6 °C

Release of options (Administrator)

102 Parameter sets

103 Measurement recorder

104 Logbook

106 Software update

107 AuditTrail

Active

Inactive

Inactive

Inactive

Active

Inactive

Abort

Select System control using arrow keys, confirm with enter.

Then select Release of options using arrow keys, confirm with

enter.

Release of options

Select the additional function to be released.

Set option to “active”. Enter the

TAN at the prompt. (Note: The TAN is only valid for the device with the corresponding serial number, see previous page.) The option is available after the TAN has been entered.

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 49 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

Signature entry

(User name, passcode)

Signature entry (user name, passcode) to open a menu

To protect the data against unauthorized query or modification, access data must be entered before a main menu from the menu selection is opened.

Two-stage security is provided by input of user name and passcode.

Menu selection

7.00 pH

25.6 °C

Enter signature:

Select:

User: ADMIN - - -

Passcode:

• • •

[enter]

Return to meas Lingua

The passcode is not visible. It is only displayed as 4 dots. The administrator can delete each user’s passcode (with entry in

AuditTrail log). The user is identified by his/her access data.

After entry of the valid signature all menus are available with the defined user rights. Only in measuring mode, the rights will be erased again. This allows accessing several menus without having to enter the access data each time.

When an invalid combination of user name and passcode is entered, access is denied with the text “Wrong passcode”. The entry “Wrong passcode” with the entered (wrong) user ID is recorded in the AuditTrail log.

Default setting for signature

Signature

User: ADMIN

Passcode: 1989

49

SW700_ENG_Version6.QXD 01.08.2005 14:08 Uhr Seite 50 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

AuditTrail: Parameter setting

Adapt function: Parameter setting/System control/AuditTrail

Parameter setting

In the System Control menu the Administrator can adjust the AuditTrail system to the requirements of his company. For example, he/she can select the protection against unauthorized query and the coding of the data in the

AuditTrail log.

AuditTrail (Administrator)

User management

AutoComplete

Access diag menu

Record AuditTrail

Coding Log

On

On

On

7.20 pH

25.6 °C

Off

Free

Off

Off

Return

• “AutoComplete” is an input assistance for the user name as known from some

Windows programs.

• Access to the Diagnostics menu can be

“free” or require input of a “Signature” according to the access authorization granted the User Management menu.

• For the recording on SmartMedia card, a coding can be activated so that the data cannot be read or queried without corresponding programs.

User management

An administrator and up to 5 users can be entered in the M 700. Each user has a user name (max. 8 characters) and a passcode (4 characters). Each user can be granted access and usage rights for each main menu level. The combination of user name and passcode is the user’s “Electronic Signature”.

Every logged modification is signed and can be clearly traced back. Different access rights for each menu can be defined for each user.

50

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:09 Uhr Seite 51 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

AuditTrail: User management

Create, edit, erase user data

Menu Display AuditTrail: User management

Menu selection

Select:

Return to meas

7.0 pH

17.1 °C

[enter]

Lingua

7.20 pH

25.6 °C

User management (Administrator)

Admin: ADMIN

User 1: BERT

User 2: ERNIE

User 3: KERMIT

User 4:

User 5:

(adj, maint, adm, diag)

(cal; maint; opl, diag)

(view)

(view, diag)

()

()

Return

7.20 pH

25.6 °C

User 1: Setup (Administrator)

User name

Passcode cal level maint level par level diag level

BERT - - - -

••••

Check

Maintenance

Operator

Diagnostics

Return

Menu selection

From the measuring mode:

Press menu key to select menu.

Select parameter setting using arrow keys, confirm with enter.

Signature for first access:

User: ADMIN

Passcode as delivered: 1989

Select:

Parameter setting/System control/

AuditTrail/User management.

Access data and rights can be created for the administrator and up to 5 users.

The administrator has always administrator rights for parameter setting.

The user enters his/her passcode after being asked to do so by the administrator. The passcode is only displayed as a series of 4 dots.

Access rights for the main menu levels:

• Calibration:

• Maintenance:

[ No access | Check | Adjustment ]

[ No access | Maintenance ]

• Parameter setting: [ No access | Viewing | Operator | Administrator ]

• Diagnostics: [ No access | Diagnostics ]

51

SW700_ENG_Version6.QXD 01.08.2005 14:09 Uhr Seite 52 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

Record AuditTrail

Functions

AutoComplete

The “AutoComplete” function facilitates the input of the user name. You only have to enter as many letters as the M 700 requires to identify the name.

This function can be switched off in the AuditTrail menu because the automatic display of the user name reduces system security, which then might not be sufficient for FDA documentation.

Access to diag menu

To protect all data against unauthorized query, it is also possible to protect the diagnostics level with a passcode (this will also affect the Favorites menu). As delivered, access to the diagnostics level is free.

Record AuditTrail

The AuditTrail log is an additional - protected if required - recording of all data on the AuditTrail card (SmartMedia card). The log is recorded in the currently selected language.

The data can be coded (to protect against unauthorized query) and secured with a checksum (to prove the authenticity of the data). The coding can be switched on or off in the AuditTrail menu of the Administrator level.

A four-digit line numbering (0000 ... 9999 P 0000 ..., cyclic) ensures uninterrupted recording. The AuditTrail log cannot be deleted by the M 700.

A buffer storage is provided to store the recorded data while the AuditTrail card is being replaced, for example. An error message “Fail AuditTrail card” is generated when the buffer is full. The oldest entries in the buffer will be recorded. They are then missing in the log (the corresponding line numbers will be missing).

When an empty AuditTrail card is inserted, the buffer storage and the current entries are transferred to the empty card.

52

M 700(X)

SW700_ENG_Version6.QXD 01.08.2005 14:09 Uhr Seite 53 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

Record AuditTrail

Parameters

The AuditTrail log cannot be queried by the M 700. It can only be evaluated using the suitable PC program. According to FDA 21 CFR Part 11 the user must ensure the transport of the data from the M 700 to evaluation and further processing or archiving. Furthermore, the user must ensure that the log data cannot be deleted from the memory card. The PC evaluation program is able to decode the data and verify the security parameters and thus to prove the authenticity of the data on the card. It allows printing the encoded data and exporting them to other PC applications.

Logbook

The logbook is provided in addition. In AuditTrail systems it runs simultaneously with the audit trail log in contrast to which it can be queried provided that the diagnostics menu is accessible. The administrator is permitted to delete the logbook data since all security-relevant data are stored in the

AuditTrail log.

Recordings in the AuditTrail log

Caution!

Data loss (even total destruction) if the card is not closed before it is pulled out. (Maintenance menu, “Close card”)

The following entries and messages are recorded:

• Measuring point, device description, module equipment and serial numbers

• Menu system log-ins

• Movements within menus (menu headlines)

• Modified parameters at the press of enter

• Calibration record at the end of a calibration or adjustment incl. user name, sensor type and serial number

• Messages such as failure and maintenance request

53

SW700_ENG_Version6.QXD 01.08.2005 14:09 Uhr Seite 54 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

AuditTrail log

Measurement point

FRONT M 700-011

BASE M 700-021 pH 2700

O2 4700

BIGBLUE vessel

0006123

0006458

0007221

0006045

6653 07/01/04 10:43:02Login: Admin (ADMIN)

6654 07.01.04

10:43:02Login: > Parameter setting (Administrator)

6655 07.01.04

10:43:02Login: > [1] Module pH 2700 (Administrator)

6656 07.01.04

10:43:02Login: < Parameter setting (Administrator)

6657 07.01.04

10:43:02Login: < Menu selection

6658 07.01.04

10:43:02Login: Admin

6659 07.01.04

10:43:02Login: > Calibration

6660 07.01.04

10:43:02Login: > [1] Module pH 2700

6661 07.01.04

10:43:02Login: > [1] Calimatic

6662 07.01.04

10:43:02Login: Warn Identical Buf

6663 07.01.04

10:43:02Login: > [1] Calimatic

6664 07.01.04

10:43:02Login: [1] Cal record

6665

6666

6667

6668

6669

Last calibration: 20.11.03 10:44

User: ADMIN

Sensor type: El 204

Serial number: 0815

Cal mode: Calimatic

6670

6671

6672

6673

Zero: 7.02 pH

Slope: 58.0 mV/pH

Impedance glass (25°C): 825.9 MOhm

Impedance ref (25°C): 119.4 kOhm

6674 0

6675

6676

6677

1st nominal buffer value: 7.00 pH

Desired value: 6.97 pH

Actual value: 7.10 pH

Delta value: 0.13 pH

6678 0

6679 0

Electrode potential: 1 mV

Cal temperature: 25.0 °C

6680 0 Response time: 19 s

6681 07.01.04

10:43:02Login: < [1] Module pH 2700

6682 07.01.04

10:43:02Login: < Calibration

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

OK

Typical AuditTrail log

(gray: logged measurement deviations)

SW700_ENG_Version6.QXD 01.08.2005 14:09 Uhr Seite 55 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

SW700_ENG_Version6.QXD 01.08.2005 14:09 Uhr Seite 56 gro Gronem:Desktop Folder:SW_6:M700_Optionen_MT_S

Additional functions

Provided by: device-specific TAN

To release: Parameter setting / System control / Release of options

Passcode Administrator level 1989 (new: ...............)

52121198 KI recorder ..........................................................7

52121199 Buffer sets to be entered (pH) ...........................13

52121200 Adaptive cal timer (pH)......................................13

52121201 ServiceScope (pH)..............................................14

52121202 Tolerance band recorder (pH).............................15

52121203 Current characteristic definable .........................17

52121204 TC ultrapure water (Cond) ................................19

52121205 Concentration determination (Cond) .................20

52121250 High CO

2 compensation (O

2

) ............................26

52121274 ISFET for ISM module pH 2700i(X).....................28

Additional functions

Provided by: device-specific TAN

Inserting the SmartMedia card..............................................30

Removing the SmartMedia card............................................30

Usage ...................................................................................31

52121192 5 loadable parameter sets .................................35

52121193 Measurement recorder ......................................37

52121194 Extended logbook .............................................40

52121195 Software update ...............................................43

52121241 AuditTrail (FDA 21 CFR Part 11).........................45

advertisement

Key Features

  • Expandable modular process analysis system
  • Device-specific additional functions to enhance capabilities
  • Data logging for comprehensive process monitoring
  • Tolerance band monitoring for quality control
  • Concentration determination for precise measurements
  • High CO2 compensation for accurate O2 measurement
  • ISFET for ISM module pH 2700i(X) for reliable pH measurement
  • 5 loadable parameter sets for flexible configuration
  • Measurement recorder for data analysis and trending
  • Extended logbook for detailed record-keeping

Related manuals

Frequently Answers and Questions

How do I activate an additional function?
You need a TAN (transaction number) to release the additional function. Contact the manufacturer with your device's serial number and hardware version to obtain the TAN.
What is the benefit of using the Tolerance band recorder?
It allows you to monitor your process within defined limits and trigger alarms if the values fall outside the tolerance band.
Can I use the M 700(X) to measure pH and conductivity simultaneously?
Yes, the modular design enables you to combine different measurement modules to meet your specific application requirements.