89601X VXA Measurement Application User`s and


Add to my manuals
1003 Pages

advertisement

89601X VXA Measurement Application User`s and | Manualzz

Agilent X-Series

Signal Analyzer

This manual provides documentation for the following X-Series Analyzers:

PXA Signal Analyzer N9030A

MXA Signal Analyzer N9020A

EXA Signal Analyzer N9010A

CXA Signal Analyzer N9000A

89601X VXA Measurement

Application User’s and

Programmer’s Reference

Option 205 Basic VSA-lite

Option 333 Connectivity to X-Series Analyzers

Option AYA Vector Modulation Analysis

Option B7R WLAN Modululation Analysis

Agilent Technologies

Notices

© Agilent Technologies, Inc. 2008-2010

No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by

United States and international copyright laws.

Trademark

Acknowledgements

Microsoft

®

is a U.S. registered trademark of Microsoft Corporation.

Windows

®

and MS Windows

®

are

U.S. registered trademarks of

Microsoft Corporation.

Adobe Reader

®

is a U.S. registered trademark of Adobe System

Incorporated.

Java™ is a U.S. trademark of Sun

Microsystems, Inc.

MATLAB

®

is a U.S. registered trademark of Math Works, Inc.

Norton Ghost™ is a U.S. trademark of Symantec Corporation.

Manual Part Number

89601-90011

Supersedes: 89601-90011 Oct. 2009

Print Date

March 2010

Printed in USA

Agilent Technologies, Inc.

1400 Fountaingrove Parkway

Santa Rosa, CA 95403

Warranty

The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein.

Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.

licensed as “Commercial computer software” as defined in DFAR 252.227-7014

(June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR

52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD

Departments and Agencies of the U.S.

Government will receive no greater than

Restricted Rights as defined in FAR

52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than

Limited Rights as defined in FAR 52.227-

14 (June 1987) or DFAR 252.227-7015

(b)(2) (November 1995), as applicable in any technical data.

Safety Notices

CAUTION

A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

WARNING

Technology Licenses

The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a

WARNING notice until the indicated conditions are fully understood and met.

Restricted Rights Legend

If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and

2

Warranty

This Agilent technologies instrument product is warranted against defects in material and workmanship for a period of one year from the date of shipment. during the warranty period, Agilent Technologies will, at its option, either repair or replace products that prove to be defective.

For warranty service or repair, this product must be returned to a service facility designated by Agilent

Technologies. Buyer shall prepay shipping charges to Agilent Technologies shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from another country.

Where to Find the Latest Information

Documentation is updated periodically. For the latest information about this analyzer, including firmware upgrades, application information, and product information, see the following URLs: http://www.agilent.com/find/pxa http://www.agilent.com/find/mxa http://www.agilent.com/find/exa http://www.agilent.com/find/cxa

To receive the latest updates by email, subscribe to Agilent Email Updates: http://www.agilent.com/find/emailupdates

Information on preventing analyzer damage can be found at: http://www.agilent.com/find/tips

Is your product software up-to-date?

Periodically, Agilent releases software updates to fix known defects and incorporate product enhancements. To search for software updates for your product, go to the Agilent Technical Support website at www.agilent.com/find/techsupport .

3

4

Contents

1. Using Help

Locating Other Help Resources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

Viewing Help on a separate Computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65

Copying the HTML Help (CHM) Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65

Copying the Acrobat (PDF) Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66

How Help is Organized . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

Help Contents Listing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

System Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

Key Descriptions for Each Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69

Key Information for Softkeys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69

Common Measurement Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70

Front Panel Keys used by the Help System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71

Navigating Windows HTML Help (CHM) Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72

HTML Help Window Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72

Basic Help Window Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73

Navigating the Help Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74

Navigating Acrobat (PDF) Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80

Adobe Reader Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80

Printing Acrobat Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81

Terms Used in This Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

Terms used in Key Parameter Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

Context Sensitive Help not Available . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85

Finding a Topic without a Mouse and Keyboard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85

Selecting a Hyperlink without a Mouse. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86

2. Additional Documentation

Documentation Categories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88

For All Measurement Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89

For Specific Measurement Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90

N6149A-2FP iDEN/WiDEN/MotoTalk Measurement Application. . . . . . . . . . . . . . . . . . . . 90

N6153A-2FP DVB-T/H Measurement Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91

N6155A-2FP ISDB-T Measurement Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91

N6156A-2FP DTMB Measurement Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91

N6158A-2FP CMMB Measurement Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91

N9020A/N9010A IQ Analyzer Measurement Application . . . . . . . . . . . . . . . . . . . . . . . . . . 91

N9020A/N9010A Spectrum Analyzer Measurement Application . . . . . . . . . . . . . . . . . . . . 91

N9061A Remote Language Compatibility Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92

N9063A Analog Demod Measurement Application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92

N9068A Phase Noise Measurement Application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92

N9069A Noise Figure Measurement Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92

N9071A GSM-EDGE Measurement Application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92

N9072A cdma2000 Measurement Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93

N9073A W-CDMA Measurement Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93

N9074A-XFP Single Acquisition Combined Fixed WiMAX Measurement Application . . . 93

N9075A 802.16 OFDMA (WiMAX/WiBro) Measurement Application . . . . . . . . . . . . . . . . . 93

N9076A 1xEV-DO Measurement Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93

N9077A-XFP Single Acquisition Combined WLAN Measurement Application . . . . . . . . . 93

N9079A TD-SCDMA Measurement Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94

N9080A LTE FDD Measurement Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94

5

Contents

N9082A LTE TDD Measurement Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .94

89601X Vector Signal Analyzer Measurement Application . . . . . . . . . . . . . . . . . . . . . . . . . .94

Application Notes & Other Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .95

3. About the Analyzer

Installing Application Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .98

Viewing a License Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .98

Obtaining and Installing a License Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .98

Missing and Old Measurement Application Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .99

X-Series Options and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100

Front-Panel Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .105

Overview of key types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .108

Display Annotations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .109

Rear-Panel Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .111

Window Control Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .114

Multi-Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .114

Zoom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .114

Next Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .115

Mouse and Keyboard Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .116

Right-Click . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .116

PC Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .117

Instrument Security & Memory Volatility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .121

4. About the Agilent 89601X VXA Signal Analyzer Measurement

What Does the Agilent 89601X VXA Signal Analyzer Measurement Application Do? . . . . .124

Installing Application Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .126

Viewing a License Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .126

Obtaining and Installing a License Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .126

Missing and Old Measurement Application Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . .127

5. System Functions

File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .130

File Explorer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .130

Page Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .131

Print . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .132

Exit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .132

Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .133

Mode Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .133

Restore Mode Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .134

*RST (Remote Command Only). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .135

Print . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .136

Quick Save . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .137

Recall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .139

State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .139

Trace (+State) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .143

Data (Import) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .146

Open… . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .150

File Open Dialog and Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .150

6

Contents

Save . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155

State. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155

Trace (+State) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158

Data (Export). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 162

Screen Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166

Save As . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170

Mass Storage Catalog (Remote Command Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173

Mass Storage Change Directory (Remote Command Only) . . . . . . . . . . . . . . . . . . . . . . . . 173

Mass Storage Copy (Remote Command Only). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173

Mass Storage Delete (Remote Command Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174

Mass Storage Data (Remote Command Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174

Mass Storage Make Directory (Remote Command Only) . . . . . . . . . . . . . . . . . . . . . . . . . 175

Mass Storage Move (Remote Command Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175

Mass Storage Remove Directory (Remote Command Only) . . . . . . . . . . . . . . . . . . . . . . . . 175

System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176

Show. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176

Power On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184

Alignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193

I/O Config . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217

Restore Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 260

Control Panel… . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264

Licensing… . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 265

Security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 268

Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 269

Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 272

System Remote Commands (Remote Commands Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . 272

User Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 275

User Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 275

User Preset All Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 276

Save User Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 277

6. Programming the Analyzer

What Programming Information is Available? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 280

IEEE Common GPIB Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281

Calibration Query . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281

Clear Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281

Standard Event Status Enable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281

Standard Event Status Register Query . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 282

Identification Query . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 282

Instrument Model Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 283

Operation Complete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 283

Query Instrument Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 283

Recall Instrument State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284

Save Instrument State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284

Service Request Enable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285

Status Byte Query . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285

Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285

Self Test Query . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 286

Wait-to-Continue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 286

7

Contents

7. Vector Analysis

AMPTD Y Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .289

Auto Couple . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .290

BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .291

Cont (Continuous) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .292

FREQ Channel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .293

Input/Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .294

Marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .295

Marker Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .296

Marker To. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .297

Meas (Measure) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .298

Meas Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .299

Avg Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .299

Average Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .299

Average Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .299

Meas Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .299

Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .300

Mode Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .301

Peak Search . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .302

Single . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303

Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .304

SPAN X Scale. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .305

Sweep/Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .306

Trace/Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .307

Select Trace. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .307

Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .307

Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .308

Digital Demod Trace Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .308

Copy to Data Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .308

Phase/Delay Properties . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .308

ACP Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .309

OBW Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .309

Trace Indicator Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .309

Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .310

View/Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .311

Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .311

Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .311

Preset View: Spectrum/Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .311

Preset View: Statistics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .312

8. Analog Demod

AMPTD Y Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .316

Auto Couple . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .317

BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .318

Cont (Continuous) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .319

FREQ Channel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .320

Input/Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .321

Marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .322

Marker Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .323

8

Contents

Marker To . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 324

Meas (Measure) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 325

Meas Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 326

Avg Number. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 326

Average Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 326

Average Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 326

Demod Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 326

Meas Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 329

Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 330

Mode Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 331

Single . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 332

Trace/Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 333

Select Trace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 333

Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 333

Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 335

Digital Demod Trace Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 335

Copy to Data Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 335

Phase/Delay Properties. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 335

ACP Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 335

OBW Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 335

Trace Indicator Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 335

Trigger. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 336

View/Display (View Presets) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 337

Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 337

Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 337

Preset View: Demod Spectrum/Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 337

Preset View: Statistics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 337

9. Digital Demod

AMPTD Y Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 343

Auto Couple . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 344

BW. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 345

Cont (Continuous) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 346

FFT Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 346

FREQ Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 347

Input/Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 348

Marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 349

Marker Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 350

Marker To . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 351

Meas (Measure) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 352

Meas Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 353

Avg Number. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 353

Average Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 353

Average Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 353

Demod Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 353

Adaptive Equalizer Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 370

Preset to Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 374

Meas Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 375

Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 376

9

Contents

Mode Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .377

Peak Search . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .378

Single . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .379

Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .380

SPAN X Scale. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .381

Sweep/Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .382

Trace/Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .383

Select Trace. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .383

Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .383

Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .391

Digital Demod Trace Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .391

Copy to Data Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .392

Phase/Delay Properties . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .392

ACP Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .392

OBW Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .392

Trace Indicator Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .392

Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .393

View/Display (View Presets) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .394

Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .394

Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .394

Preset View: Demod Quad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .394

10. WLAN DSSS

AMPTD Y Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .401

Auto Couple . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .402

BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .403

FFT Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .403

Cont (Continuous) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .404

FREQ Channel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .405

Input/Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .406

Marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .407

Marker Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .408

Marker To. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .409

Meas (Measure) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .410

Meas Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .411

Demod Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .411

Preset to Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .415

Meas Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .415

Advanced. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .419

Meas Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .423

Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .424

Mode Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .425

Peak Search . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .426

Single . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .427

Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .428

SPAN X Scale. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .429

Sweep/Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .430

Trace/Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .431

Select Trace. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .431

10

Contents

Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 431

Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 438

Digital Demod Trace Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 438

Copy to Data Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 438

Phase/Delay Properties. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 438

ACP Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 438

OBW Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 438

Trace Indicator Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 438

Trigger. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 439

View/Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 440

Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 440

Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 440

Preset View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 440

11. WLAN OFDM

AMPTD Y Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 448

Auto Couple . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 449

BW. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 450

FFT Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 450

Cont (Continuous) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 451

FREQ Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 452

Input/Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 453

Marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 454

Marker Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 455

Marker To . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 456

Meas (Measure) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 457

Meas Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 458

Avg Number. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 458

Average Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 458

Average Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 458

Demod Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 459

Preset to Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 461

Meas Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 463

Subcarriers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 466

Advanced . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 467

Meas Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 475

Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 476

Mode Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 477

Peak Search . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 478

Recall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 479

State. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 479

Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 479

Restart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 480

Save . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 481

State. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 481

Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 481

Screen Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 481

Single . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 482

Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 483

11

Contents

SPAN X Scale. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .484

Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .484

Full Span. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .484

Select Trace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .484

X Scale. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .485

Freq Annotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .485

All Frequency Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .485

Copy X Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .485

Sweep/Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .486

Trace/Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .487

Select Trace. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .487

Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .487

Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .494

Digital Demod Trace Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .494

Copy to Data Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .494

Phase/Delay Properties . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .494

ACP Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .495

OBW Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .495

Trace Indicator Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .495

Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .496

View/Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .497

Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .497

Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .497

Preset View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .497

12. Common Measurement Functions

Common Measurement Functions 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1

AMPTD Y Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1

Auto Couple . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33

Cont (Continuous Measurement/Sweep). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35

FREQChannel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36

Input/Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43

Marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113

Marker Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .132

Marker To . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .151

Meas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .155

Meas Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .168

Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .213

Mode Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .229

Peak Search . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .241

Recall. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .255

Restart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .258

Save . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .260

Single (Single Measurement/Sweep) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .263

Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .263

Sweep/Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .263

Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .290

View/Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .349

Common Measurement Functions 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .361

12

Contents

AMPTD Y Scale (Amplitude) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 361

BW (Bandwidth) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 368

FREQ Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 371

Marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 374

Marker -> (Marker To) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 390

Marker Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 392

Meas Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 400

Peak Search . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 406

Recall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 410

Save . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 413

SPAN X Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 416

Sweep/Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 424

Trace/Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 428

Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 470

View/Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 479

Remote SCPI Commands and Data Queries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 482

13

Contents

14

List of Commands

*CAL? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201

*CLS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281

*ESE <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281

*ESE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281

*ESR? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 282

*IDN?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 283

*OPC? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 283

*OPC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 283

*OPT? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284

*RCL <register #>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284

*RST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135

*SAV <register #> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284

*SRE <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285

*SRE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285

*STB?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285

*TRG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285

*TST? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 286

*WAI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 286

:ABORt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 289

:CALCulate:<meas>:DATA[1]|2|3|4:HEADer:NAMes? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 493

:CALCulate:<meas>:DATA[1]|2|3|4:HEADer:STRing? <string> . . . . . . . . . . . . . . . . . . . . . . . . . 494

:CALCulate:<meas>:DATA[1]|2|3|4:HEADer:TYPE? <string> . . . . . . . . . . . . . . . . . . . . . . . . . . 493

:CALCulate:<meas>:DATA[1]|2|3|4:HEADer[:NUMBer]? <string>. . . . . . . . . . . . . . . . . . . . . . . 494

:CALCulate:<meas>:DATA[1]|2|3|4:NAMes? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 431

:CALCulate:<meas>:DATA[1]|2|3|4:POINts? [OFF|ON|0|1] . . . . . . . . . . . . . . . . . . . . . . . . . . . 489

:CALCulate:<meas>:DATA[1]|2|3|4:RAW:COMPlex? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 488

:CALCulate:<meas>:DATA[1]|2|3|4:RAW:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 489

:CALCulate:<meas>:DATA[1]|2|3|4:RAW? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 488

:CALCulate:<meas>:DATA[1]|2|3|4:TABLe:NAMes? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 491

:CALCulate:<meas>:DATA[1]|2|3|4:TABLe:STRing? [<string>] . . . . . . . . . . . . . . . . . . . . . . . . . 490

:CALCulate:<meas>:DATA[1]|2|3|4:TABLe:UNIT? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 491

15

List of Commands

:CALCulate:<meas>:DATA[1]|2|3|4:TABLe[:NUMBer]? [<string>] . . . . . . . . . . . . . . . . . . . . . . .490

:CALCulate:<meas>:DATA[1]|2|3|4? [Y|X|XY[,OFF|ON|0|1]] . . . . . . . . . . . . . . . . . . . . . . . . .487

:CALCulate:<meas>:MARKer:AOFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .390

:CALCulate:<meas>:MARKer:COUPle[:STATe] OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . .389

:CALCulate:<meas>:MARKer:COUPle[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .389

:CALCulate:<meas>:MARKer:TABLe[:STATe] OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .388

:CALCulate:<meas>:MARKer:TABLe[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .388

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:CFORmat RECTangular|POLar . .

386

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:CFORmat? . . . . . . . . . . . . . . . . .386

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:CPSearch[:STATe] ON|OFF|1|0 .

409

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:CPSearch[:STATe]? . . . . . . . . . .409

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt:X? . . . . . . . . . . . . . . . . .388

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt[:STATe] OFF|ON|0|1. . .

387

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt[:STATe]? . . . . . . . . . . .387

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion BPOWer|BDENsity|=OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .393

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BAND:CENTer <real> .

397

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BAND:CENTer? . . .397

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BAND:LEFT <real> 398

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BAND:LEFT? . . . . .398

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BAND:RIGHt <real> . .

399

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BAND:RIGHt?. . . . .399

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BAND:SPAN <real>398

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BAND:SPAN? . . . . .398

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BDENsity:CTYPe

MEAN|RMS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .396

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BDENsity:CTYPe? .396

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BPOWer:CTYPe

MEAN|RMS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .395

16

List of Commands

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BPOWer:CTYPe? . . 395

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion? . . . . . . . . . . . . . . . . 393

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MAXimum . . . . . . . . . . . . . . . . 406

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MAXimum:LEFT . . . . . . . . . . . 408

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MAXimum:NEXT. . . . . . . . . . . 407

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MAXimum:PREVious . . . . . . . 407

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MAXimum:RIGHt . . . . . . . . . . 408

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MINimum . . . . . . . . . . . . . . . . . 410

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MODE POSition|DEL-

Ta|FIXed|=OFF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 376

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MODE? . . . . . . . . . . . . . . . . . . . 376

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:REFerence <integer> . . . . . . . . 385

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:REFerence?. . . . . . . . . . . . . . . . 385

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:TRACe <integer> . . . . . . . . . . . 386

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:TRACe? . . . . . . . . . . . . . . . . . . . 386

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X <real> . . . . . . . . . . . . . . . . . . 380

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X:UNIT? . . . . . . . . . . . . . . . . . . 381

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X? . . . . . . . . . . . . . . . . . . . . . . . 380

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Y:IMAGinary <real> . . . . . . . . 383

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Y:IMAGinary?. . . . . . . . . . . . . . 383

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Y:UNIT? . . . . . . . . . . . . . . . . . . 382

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Y[:REAL] <real> . . . . . . . . . . . 382

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Y[:REAL]?. . . . . . . . . . . . . . . . . 382

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Z <real> . . . . . . . . . . . . . . . . . . 384

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Z:UNIT? . . . . . . . . . . . . . . . . . . 384

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Z?. . . . . . . . . . . . . . . . . . . . . . . . 384

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:X]:POSition <real> . . . . . . . . . 381

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:X]:POSition? . . . . . . . . . . . . . . 381

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:BANDwidth|BWIDth:INTegration

<bandwidth> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 457

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:BANDwidth|BWIDth:INTegration?. 457

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:FILTer:RRC:ALPHa <real>. . . . . . . . 458

17

List of Commands

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:FILTer:RRC:ALPHa? . . . . . . . . . . . . .458

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:FILTer:RRC:STATe OFF|ON|0|1 . .458

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:FILTer:RRC:STATe? . . . . . . . . . . . . . .458

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:FREQuency <freq> . . . . . . . . . . . . . . .457

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:FREQuency? . . . . . . . . . . . . . . . . . . . .457

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:FILTer:RRC:STATe OFF|ON|0|1 . . .462

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:FILTer:RRC:STATe? . . . . . . . . . . . . . . .462

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:BANDwidth|BWIDth:INTegration

<bandwidth>,… . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .460

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:BANDwidth|BWIDth:INTegration?

460

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:FILTer:RRC:ALPHa <real>,… . .461

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:FILTer:RRC:ALPHa?. . . . . . . . . .461

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:FREQuency <freq>,… . . . . . . . . .459

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:FREQuency? . . . . . . . . . . . . . . . .459

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:RCARrier <reall>,… . . . . . . . . . .462

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:RCARrier:TEST OFF|ON|0|1,… . .

462

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:RCARrier:TEST? . . . . . . . . . . . . .462

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:RCARrier? . . . . . . . . . . . . . . . . . .462

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:STATe OFF|ON|0|1,… . . . . . . .459

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:STATe? . . . . . . . . . . . . . . . . . . . . .459

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . .456

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:STATe?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .456

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:CENTroid?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .467

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:LIMit:FBLimit <freq> . . . . . . . . . . . . . . . . . . . .467

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:LIMit:FBLimit?. . . . . . . . . . . . . . . . . . . . . . . . . .467

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:LIMit[:TEST] OFF|ON|0|1 . . . . . . . . . . . . . . .467

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:LIMit[:TEST]? . . . . . . . . . . . . . . . . . . . . . . . . . . .467

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:PERCent <real> . . . . . . . . . . . . . . . . . . . . . . . . .466

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:PERCent? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .466

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . .466

18

List of Commands

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 466

:CALCulate:BWIDth|BANDwidth:NDB <rel_ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193

:CALCulate:BWIDth|BANDwidth:NDB? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193

:CALCulate:BWIDth|BANDwidth:RESult? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193

:CALCulate:BWIDth|BANDwidth[:STATe] OFF|ON|0|1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193

:CALCulate:BWIDth|BANDwidth[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193

:CALCulate:CLIMits:FAIL? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159

:CALCulate:DATA:REGister:ALL:REMove . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 451

:CALCulate:DATA:REGister[1]|2|3|4|5|6:EMPTy? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 450

:CALCulate:DATA:REGister[1]|2|3|4|5|6:REMove . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 450

:CALCulate:DATA<n>:COMPress? BLOCk|CFIT|MAXimum|MINimum|MEAN|DMEan|RMS|RMSCubed|SAMPle|SDEViation|PPHase [,<soffset>[,<length>[,<roffset>[,<rlimit>]]]] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160

:CALCulate:DATA[1]|2|3|4|5|6:PEAKs? <threshold>,<excursion>[,AMPLitude|FREQuency|TIME[,ALL|GTDLine|LTDLine]] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165

:CALCulate:DATA[1]|2|3|4|5|6:PEAKs? <threshold>,<excursion>[,AMPLitude|FREQuency|TIME]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165

:CALCulate:DATA[n]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159

:CALCulate:DDEMod:IQRotation <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 368

:CALCulate:DDEMod:IQRotation? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 368

:CALCulate:DDEMod:NORMalize OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 368

:CALCulate:DDEMod:NORMalize:? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 368

:CALCulate:DDEMod:PPSYmbol:COUPle OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 357

:CALCulate:DDEMod:PPSYmbol:COUPle? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 357

:CALCulate:LIMit[1]|2|3|4|5|6:CLEar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191

:CALCulate:LIMit[1]|2|3|4|5|6:CONTrol:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189

:CALCulate:LIMit[1]|2|3|4|5|6:CONTrol[:DATA] <x>, <x>, …. . . . . . . . . . . . . . . . . . . . . . . . . . 189

:CALCulate:LIMit[1]|2|3|4|5|6:CONTrol[:DATA]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189

:CALCulate:LIMit[1]|2|3|4|5|6:FAIL? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191

:CALCulate:LIMit[1]|2|3|4|5|6:LOWer:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190

:CALCulate:LIMit[1]|2|3|4|5|6:LOWer[:DATA] <ampl>, … . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190

:CALCulate:LIMit[1]|2|3|4|5|6:LOWer[:DATA]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190

19

List of Commands

:CALCulate:LIMit[1]|2|3|4|5|6:STATe ON|OFF|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .188

:CALCulate:LIMit[1]|2|3|4|5|6:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .188

:CALCulate:LIMit[1]|2|3|4|5|6:UPPer:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .190

:CALCulate:LIMit[1]|2|3|4|5|6:UPPer[:DATA] <ampl>, <ampl>, … . . . . . . . . . . . . . . . . . . . . . .190

:CALCulate:LIMit[1]|2|3|4|5|6:UPPer[:DATA]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .190

:CALCulate:LLINe:ALL:DELete. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .187

:CALCulate:LLINe:CMODe FIXed|RELative . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .192

:CALCulate:LLINe:CMODe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .192

:CALCulate:LLINe:CONTrol:DOMain FREQuency|TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .187

:CALCulate:LLINe:CONTrol:DOMain? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .187

:CALCulate:LLINe:TEST OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .186

:CALCulate:LLINe:TEST?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .186

:CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:CMODe:RELative ON|OFF|1|0 . . . . . . . . . . . . .179

:CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:CMODe:RELative? . . . . . . . . . . . . . . . . . . . . . . . .179

:CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:INTerpolate:TYPE LOGarithmic|LINear . . . . . .177

:CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:INTerpolate:TYPE? . . . . . . . . . . . . . . . . . . . . . . . .177

:CALCulate:LLINe[1]|2|3|4|5|6:BUILd

TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .183

:CALCulate:LLINe[1]|2|3|4|5|6:COMMent "text". . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .180

:CALCulate:LLINe[1]|2|3|4|5|6:COMMent? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .180

:CALCulate:LLINe[1]|2|3|4|5|6:CONTrol:INTerpolate:TYPE LOGarithmic|LINear . . . . . . . .177

:CALCulate:LLINe[1]|2|3|4|5|6:CONTrol:INTerpolate:TYPE? . . . . . . . . . . . . . . . . . . . . . . . . . .177

:CALCulate:LLINe[1]|2|3|4|5|6:COPY LLINE1|LLINE2|LLINE3|LLINE4|LLINE5|LLINE6 .

183

:CALCulate:LLINe[1]|2|3|4|5|6:DATA <x>,<ampl>,<connect> . . . . . . . . . . . . . . . . . . . . . . . . . .188

:CALCulate:LLINe[1]|2|3|4|5|6:DATA:MERGe <x-axis>,<ampl>,<connected> . . . . . . . . . . . . .192

:CALCulate:LLINe[1]|2|3|4|5|6:DATA?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .188

:CALCulate:LLINe[1]|2|3|4|5|6:DELete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .185

:CALCulate:LLINe[1]|2|3|4|5|6:DESCription “Description” . . . . . . . . . . . . . . . . . . . . . . . . . . . .179

:CALCulate:LLINe[1]|2|3|4|5|6:DESCription? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .179

:CALCulate:LLINe[1]|2|3|4|5|6:DISPlay OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .174

:CALCulate:LLINe[1]|2|3|4|5|6:DISPlay? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .174

20

List of Commands

:CALCulate:LLINe[1]|2|3|4|5|6:FAIL? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188

:CALCulate:LLINe[1]|2|3|4|5|6:FREQuency:CMODe:RELative ON|OFF|1|0 . . . . . . . . . . . . 178

:CALCulate:LLINe[1]|2|3|4|5|6:FREQuency:CMODe:RELative? . . . . . . . . . . . . . . . . . . . . . . . 178

:CALCulate:LLINe[1]|2|3|4|5|6:MARGin <rel_ampl>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180

:CALCulate:LLINe[1]|2|3|4|5|6:MARGin:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . 180

:CALCulate:LLINe[1]|2|3|4|5|6:MARGin:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180

:CALCulate:LLINe[1]|2|3|4|5|6:MARGin? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180

:CALCulate:LLINe[1]|2|3|4|5|6:OFFSet:UPDate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185

:CALCulate:LLINe[1]|2|3|4|5|6:OFFSet:X <value> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184

:CALCulate:LLINe[1]|2|3|4|5|6:OFFSet:X? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184

:CALCulate:LLINe[1]|2|3|4|5|6:OFFSet:Y <rel ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184

:CALCulate:LLINe[1]|2|3|4|5|6:OFFSet:Y? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184

:CALCulate:LLINe[1]|2|3|4|5|6:TRACe 1|2|3|4|5|6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175

:CALCulate:LLINe[1]|2|3|4|5|6:TRACe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175

:CALCulate:LLINe[1]|2|3|4|5|6:TYPE UPPer|LOWer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176

:CALCulate:LLINe[1]|2|3|4|5|6:TYPE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176

:CALCulate:MAMarker:COUPling ON|OFF|1|0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150

:CALCulate:MAMarker:COUPling? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150

:CALCulate:MAMarker:DETector[1]|2|3 OFF|NORMal|AVERage|POSitive|SAMPle|NEGative|QPEak|EAVerage|RAVerage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146

:CALCulate:MAMarker:DETector[1]|2|3:DWELl <dwell time> . . . . . . . . . . . . . . . . . . . . . . . . . . 146

:CALCulate:MAMarker:DETector[1]|2|3:DWELl? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146

:CALCulate:MAMarker:DETector[1]|2|3? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146

:CALCulate:MAMarker:PCENter ON|OFF|1|0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150

:CALCulate:MAMarker:PCENter? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150

:CALCulate:MARKer:AOFF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132

:CALCulate:MARKer:COUPle[:STATe] OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131

:CALCulate:MARKer:COUPle[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131

:CALCulate:MARKer:PEAK:EXCursion <rel_ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 246

:CALCulate:MARKer:PEAK:EXCursion:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . 246

:CALCulate:MARKer:PEAK:EXCursion:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 246

21

List of Commands

:CALCulate:MARKer:PEAK:EXCursion?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .246

:CALCulate:MARKer:PEAK:SEARch:MODE MAXimum|PARameter . . . . . . . . . . . . . . . . . . . . . .244

:CALCulate:MARKer:PEAK:SEARch:MODE?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .244

:CALCulate:MARKer:PEAK:SORT FREQuency|AMPLitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . .249

:CALCulate:MARKer:PEAK:SORT? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .249

:CALCulate:MARKer:PEAK:TABLe:READout ALL|GTDLine|LTDLine . . . . . . . . . . . . . . . . . . . .250

:CALCulate:MARKer:PEAK:TABLe:READout? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .250

:CALCulate:MARKer:PEAK:TABLe:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .249

:CALCulate:MARKer:PEAK:TABLe:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .249

:CALCulate:MARKer:PEAK:THReshold <ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .247

:CALCulate:MARKer:PEAK:THReshold:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . .247

:CALCulate:MARKer:PEAK:THReshold:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .247

:CALCulate:MARKer:PEAK:THReshold? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .247

:CALCulate:MARKer:TABLe[:STATe] OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .127

:CALCulate:MARKer:TABLe[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .127

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:CPSearch[:STATe] ON|OFF|1|0 . . . . .252

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:CPSearch[:STATe]? . . . . . . . . . . . . . . . . .252

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt:GATetime <time> . . . . . . . . . . .131

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt:GATetime:AUTO OFF|ON|0|1 . .

131

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt:GATetime:AUTO? . . . . . . . . . . .131

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt:GATetime?. . . . . . . . . . . . . . . . .131

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt:X? . . . . . . . . . . . . . . . . . . . . . . . .128

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt[:STATe] OFF|ON|0|1. . . . . . .127

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt[:STATe]? . . . . . . . . . . . . . . . . . .127

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion NOISe|BPOWer|BDENsity|OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .132

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BAND:LEFT <freq> . . . . . . .139

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BAND:LEFT? . . . . . . . . . . . .139

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BAND:RIGHt <freq> . . . . . .140

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BAND:RIGHt?. . . . . . . . . . . .140

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BAND:SPAN <freq> . . . . . . .138

22

List of Commands

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:BAND:SPAN?. . . . . . . . . . . . 138

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:MAMarker? . . . . . . . . . . . . . 141

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion? . . . . . . . . . . . . . . . . . . . . . . . 132

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:LINes[:STATe] OFF|ON|0|1 . . . . . . . . 126

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:LINes[:STATe]? . . . . . . . . . . . . . . . . . . . 126

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MAXimum . . . . . . . . . . . . . . . . . . . . . . . 241

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MAXimum:LEFT . . . . . . . . . . . . . . . . . . 243

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MAXimum:NEXT. . . . . . . . . . . . . . . . . . 242

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MAXimum:RIGHt . . . . . . . . . . . . . . . . . 242

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MINimum . . . . . . . . . . . . . . . . . . . . . . . . 254

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MODE POSition|DELTa|FIXed|OFF. 114

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MODE? . . . . . . . . . . . . . . . . . . . . . . . . . . 114

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:PTPeak . . . . . . . . . . . . . . . . . . . . . . . . . . 253

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:REFerence <integer> . . . . . . . . . . . . . . 120

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:REFerence?. . . . . . . . . . . . . . . . . . . . . . . 120

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:TRACe 1|2|3|4|5|6. . . . . . . . . . . . . . . 124

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:TRACe:AUTO OFF|ON|0|1 . . . . . . . . 126

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:TRACe:AUTO? . . . . . . . . . . . . . . . . . . . . 126

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:TRACe? . . . . . . . . . . . . . . . . . . . . . . . . . . 124

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X <freq> . . . . . . . . . . . . . . . . . . . . . . . . . 114

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X:POSition <real> . . . . . . . . . . . . . . . . . 116

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X:POSition? . . . . . . . . . . . . . . . . . . . . . . 116

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X:READout FREQuency|TIME|ITIMe|PERiod . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X:READout:AUTO ON|OFF|1|0 . . . . 121

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X:READout:AUTO? . . . . . . . . . . . . . . . . 121

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X:READout? . . . . . . . . . . . . . . . . . . . . . . 121

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Y <real> . . . . . . . . . . . . . . . . . . . . . . . . . 116

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Y? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:CENTer . . . . . . . . . . . . . . . . . . . . 151

23

List of Commands

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:DELTa:CENTer. . . . . . . . . . . . . . .154

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:DELTa:SPAN . . . . . . . . . . . . . . . .153

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:RLEVel . . . . . . . . . . . . . . . . . . . . .154

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:STARt . . . . . . . . . . . . . . . . . . . . . .152

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:STEP . . . . . . . . . . . . . . . . . . . . . . .152

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:STOP . . . . . . . . . . . . . . . . . . . . . . .153

:CALCulate:TRACe[1]|2|3|4|5|6:FAIL? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .191

:CALCulate:W11A:NORMalize OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .473

:CALCulate:W11A:NORMalize? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .473

:CALCulate:W11B:NORMalize OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .422

:CALCulate:W11B:NORMalize? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .422

:CALibration:AUTO ON|PARTial|OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .193

:CALibration:AUTO:ALERt TTEMperature|DAY|WEEK|NONE . . . . . . . . . . . . . . . . . . . . . . . . .196

:CALibration:AUTO:ALERt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .196

:CALibration:AUTO:MODE ALL|NRF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .196

:CALibration:AUTO:MODE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .196

:CALibration:AUTO:TIME:OFF? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .206

:CALibration:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .193

:CALibration:DATA:BACKup <filename> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .214

:CALibration:DATA:DEFault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .208

:CALibration:DATA:RESTore <filename> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .214

:CALibration:FREQuency:REFerence:COARse <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .217

:CALibration:FREQuency:REFerence:COARse? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .217

:CALibration:FREQuency:REFerence:FINE <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .217

:CALibration:FREQuency:REFerence:FINE?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .217

:CALibration:FREQuency:REFerence:MODE CALibrated|USER. . . . . . . . . . . . . . . . . . . . . . . . . .216

:CALibration:FREQuency:REFerence:MODE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .216

:CALibration:IQ:FLATness:I . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .103

:CALibration:IQ:FLATness:I|IBAR|Q|QBAR:TIME? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .107

:CALibration:IQ:FLATness:IBAR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104

:CALibration:IQ:FLATness:Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .105

24

List of Commands

:CALibration:IQ:FLATness:QBAR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106

:CALibration:IQ:ISOLation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102

:CALibration:IQ:ISOLation:TIME? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102

:CALibration:IQ:PROBe:I. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109

:CALibration:IQ:PROBe:I|IBAR|Q|QBAR:TIME? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113

:CALibration:IQ:PROBe:I|Q:CLEar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

:CALibration:IQ:PROBe:IBar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110

:CALibration:IQ:PROBe:Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111

:CALibration:IQ:PROBe:QBar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112

:CALibration:NRF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201

:CALibration:NRF? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201

:CALibration:RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203

:CALibration:RF? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203

:CALibration:TEMPerature:CURRent?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204

:CALibration:TEMPerature:LALL? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205

:CALibration:TEMPerature:LPReselector? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206

:CALibration:TEMPerature:LRF? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206

:CALibration:TIME:LALL? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205

:CALibration:TIME:LPReselector? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206

:CALibration:TIME:LRF?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205

:CALibration:YTF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 215

:CALibration:YTF? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 215

:CALibration[:ALL]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200

:CALibration[:ALL]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200

:CONFigure:ADEMod:NDEFault. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 315

:CONFigure:ADEMod. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 315

:CONFigure:DDEMod:NDEFault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 339

:CONFigure:DDEMod. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 339

:CONFigure:VECTor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 299

:CONFigure:VECTor:NDEFault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 287

:CONFigure:VECTor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 287

25

List of Commands

:CONFigure:W11A:NDEFault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .445

:CONFigure:W11A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .445

:CONFigure:W11B:NDEFault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .397

:CONFigure:W11B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .397

:CONFigure?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .158

:COUPle ALL|NONE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33

:DISPlay:<meas>:AFPoints OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .422

:DISPlay:<meas>:AFPoints? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .422

:DISPlay:<meas>:FANNotation CSPan|SSTop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .423

:DISPlay:<meas>:FANNotation?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .423

:DISPlay:<meas>:TRACe[1]|2|3|4:COPY D1|D2|D3|D4|D5|D6 . . . . . . . . . . . . . . . . . . . . . . . .450

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:ALIN? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .449

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:ALINe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . .449

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:EYE:COUNt <real> . . . . . . . . . . . . . . . . . . . . . . . . .448

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:EYE:COUNt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .448

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol BARS|DOTS|OFF . . . . . . . . . . . . . . . . . .445

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol:FORMat HEXadecimal|BINary . . . . . . .447

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol:FORMat? . . . . . . . . . . . . . . . . . . . . . . . . . .447

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol:SHAPe CIRCle|CROSs|OFF. . . . . . . . . .446

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol:SHAPe? . . . . . . . . . . . . . . . . . . . . . . . . . . .446

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol:SIZE <real> . . . . . . . . . . . . . . . . . . . . . . . .446

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol:SIZE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . .446

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .445

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:UNIT:FREQuency CARRier|HZ . . . . . . . . . . . . . .449

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:UNIT:FREQuency? . . . . . . . . . . . . . . . . . . . . . . . . .449

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:UNIT:TIME SEC|SYMBol . . . . . . . . . . . . . . . . . . .448

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:UNIT:TIME? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .448

:DISPlay:<meas>:TRACe[1]|2|3|4:FEED <string> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .430

:DISPlay:<meas>:TRACe[1]|2|3|4:FEED?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .430

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat MLOG|MLINear|REAL|IMAGinary|VEC-

Tor|CONS|PHASe|UPHase|IEYE|QEYE|TRELlis|GDELay|MLGLinear . . . . . . . . . . . . . . . .444

26

List of Commands

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat:DELay:APERture <real> . . . . . . . . . . . . . . . . . . . . 453

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat:DELay:APERture? . . . . . . . . . . . . . . . . . . . . . . . . . 453

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat:PHASe:OFFSet <real> . . . . . . . . . . . . . . . . . . . . . . 452

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat:PHASe:OFFSet? . . . . . . . . . . . . . . . . . . . . . . . . . . . 452

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat:PHASe:UNWRap:REFerence <real> . . . . . . . . . . . 452

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat:PHASe:UNWRap:REFerence? . . . . . . . . . . . . . . . . 452

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 444

:DISPlay:<meas>:TRACe[1]|2|3|4:RLINe OFF|ON|0|1 :DIS-

Play:<meas>:TRACe[1]|2|3|4:RLINe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 368

:DISPlay:<meas>:TRACe[1]|2|3|4:VHCenter <real>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 367

:DISPlay:<meas>:TRACe[1]|2|3|4:VHCenter? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 367

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:COUPle OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . 420

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:COUPle? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 420

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:RLEVel <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . 420

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:RLEVel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 420

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:RPOSition LEFT|CENTer|RIGHt . . . . . . . . . . . 422

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:RPOSition? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 422

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:SPAN <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 421

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:SPAN? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 421

:DISPlay:<meas>:TRACe[1]|2|3|4:Y:LRATio <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 366

:DISPlay:<meas>:TRACe[1]|2|3|4:Y:LRATio? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 366

:DISPlay:<meas>:TRACe[1]|2|3|4:Y:UNIT:PREFerence AUTO|PEAK|RMS|POWer|MRMS 365

:DISPlay:<meas>:TRACe[1]|2|3|4:Y:UNIT:PREFerence? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 365

:DISPlay:<meas>:TRACe[1]|2|3|4:Y:UNIT? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 366

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:AUTO:ONCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . 362

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:PDIVision <real> . . . . . . . . . . . . . . . . . . . . . . . . . 363

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:PDIVision? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 363

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:RLEVel <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . 363

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:RLEVel:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . 364

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:RLEVel:AUTO?. . . . . . . . . . . . . . . . . . . . . . . . . . . 364

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:RLEVel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 363

27

List of Commands

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:RPOSition TOP|CENTer|BOTTom. . . . . . . . . . .365

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:RPOSition? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .365

:DISPlay:<meas>:WINDow:FORMat SINGle|TWO|TRI|QUAD|GR2X3|GR3X2 . . . . . . . . . . . .481

:DISPlay:<meas>:WINDow:FORMat SINGle|TWO|TRI|QUAD . . . . . . . . . . . . . . . . . . . . . . . . . .481

:DISPlay:<meas>:WINDow:FORMat? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .481

:DISPlay:<meas>:WINDow:FORMat? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .481

:DISPlay:<measurement>:ANNotation:TITLe:DATA <string> . . . . . . . . . . . . . . . . . . . . . . . . . . . .354

:DISPlay:<measurement>:ANNotation:TITLe:DATA? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .354

:DISPlay:ACTivefunc[:STATe] ON|OFF|1|0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .353

:DISPlay:ACTivefunc[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .353

:DISPlay:ADEMod:TRACe[1]|2|3|4:CARRier:FREQuency? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .329

:DISPlay:ADEMod:VIEW:PRESet DSPectrum|STATistics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .337

:DISPlay:ANNotation:MBAR[:STATe] OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .351

:DISPlay:ANNotation:MBAR[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .351

:DISPlay:ANNotation:SCReen[:STATe] OFF|ON|0|1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .352

:DISPlay:ANNotation:SCReen[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .352

:DISPlay:ANNotation:TRACe[:STATe] ON|OFF|1|0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .352

:DISPlay:ANNotation:TRACe[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .352

:DISPlay:BACKlight ON|OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .357

:DISPlay:BACKlight:INTensity <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .358

:DISPlay:BACKlight:INTensity? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .358

:DISPlay:BACKlight?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .357

:DISPlay:DDEMod:VIEW:PRESet QUAD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .394

:DISPlay:ENABle OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .359

:DISPlay:ENABle? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .359

:DISPlay:FSCReen[:STATe] OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .358

:DISPlay:FSCReen[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .358

:DISPlay:THEMe TDColor|TDMonochrome|FCOLor|FMONochrome . . . . . . . . . . . . . . . . . . . . .357

:DISPlay:THEMe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .357

:DISPlay:VECTor:VIEW:PRESet SPECtrum|STATistics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .311

:DISPlay:W11A:VIEW:PRESet BASic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .497

28

List of Commands

:DISPlay:W11A:VIEW:PRESet BASic|EVM|POWer|DIAGnostic. . . . . . . . . . . . . . . . . . . . . . . . . 497

:DISPlay:W11A:VIEW:PRESet DIAGnostic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500

:DISPlay:W11A:VIEW:PRESet EVM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 498

:DISPlay:W11A:VIEW:PRESet POWer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 499

:DISPlay:W11B:VIEW:PRESet BASic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 440

:DISPlay:W11B:VIEW:PRESet BASic|EVM|POWer|DIAGnostic. . . . . . . . . . . . . . . . . . . . . . . . . 440

:DISPlay:W11B:VIEW:PRESet DIAGnostic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 443

:DISPlay:W11B:VIEW:PRESet EVM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 441

:DISPlay:W11B:VIEW:PRESet POWer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 442

:DISPlay:WINDow:FORMat:TILE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114

:DISPlay:WINDow:FORMat:ZOOM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114

:DISPlay:WINDow:MAMarker:POSition LEFT|RIGHt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145

:DISPlay:WINDow:MAMarker:POSition?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145

:DISPlay:WINDow:MAMarker[:STATe] ON|OFF|1|0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145

:DISPlay:WINDow:MAMarker[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145

:DISPlay:WINDow[:SELect] <number> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115

:DISPlay:WINDow[:SELect]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115

:DISPlay:WINDow[1]:ANNotation[:ALL] OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 356

:DISPlay:WINDow[1]:ANNotation[:ALL]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 356

:DISPlay:WINDow[1]:TRACe:GRATicule:GRID[:STATe] OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . 355

:DISPlay:WINDow[1]:TRACe:GRATicule:GRID[:STATe]?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 355

:DISPlay:WINDow[1]:TRACe:Y:DLINe <ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 355

:DISPlay:WINDow[1]:TRACe:Y:DLINe:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 355

:DISPlay:WINDow[1]:TRACe:Y:DLINe:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 355

:DISPlay:WINDow[1]:TRACe:Y:DLINe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 355

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:PDIVision <rel_ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:PDIVision? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel:OFFSet <rel_ampl> . . . . . . . . . . . . . . . . . . . . . . 26

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel:OFFSet?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

29

List of Commands

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:SPACing LINear|LOGarithmic . . . . . . . . . . . . . . . . . . . .17

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:SPACing? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17

:FETCh:<meas>[n]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .482

:FETCh:ADEMod[n]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .315

:FETCh:DDEMod[n]?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .339

:FETCh:FCAPture? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .498

:FETCh:VECTor[n]?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .287

:FETCh:W11A[n]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .445

:FETCh:W11B[n]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .397

:FORMat:BORDer NORMal|SWAPped . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .168

:FORMat:BORDer? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .168

:FORMat[:TRACe][:DATA] ASCii|INTeger,32|REAL,32 |REAL,64 . . . . . . . . . . . . . . . . . . . . . . . .166

:FORMat[:TRACe][:DATA]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .166

:GLOBal:FREQuency:CENTer[:STATe] 1|0|ON|OFF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .228

:GLOBal:FREQuency:CENTer[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .228

:HCOPy:ABORt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .136

:HCOPy[:IMMediate] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .136

:INITiate:ADEMod. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .315

:INITiate:CONTinuous OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35

:INITiate:CONTinuous? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35

:INITiate:DDEMod. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .339

:INITiate:FCAPture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .497

:INITiate:PAUSe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .275

:INITiate:RESTart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .258

:INITiate:RESume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .275

:INITiate:VECTor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .287

:INITiate:W11A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .445

:INITiate:W11B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .397

:INITiate[:IMMediate] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .258

:INPut:COUPling AC|DC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45

:INPut:COUPling:I|Q DC|LFR1|LFR2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58

30

List of Commands

:INPut:COUPling:I|Q?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58

:INPut:COUPling? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

:INPut:IMPedance:IQ U50|B50|U1M|B1M . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

:INPut:IMPedance:IQ? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

:INPut:IMPedance:REFerence <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

:INPut:IMPedance:REFerence? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

:INPut:IQ:MIRRored OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

:INPut:IQ:MIRRored?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

:INPut:IQ:Q:DIFFerential OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

:INPut:IQ:Q:DIFFerential? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

:INPut:IQ[:I]:DIFFerential OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

:INPut:IQ[:I]:DIFFerential? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

:INPut:OFFSet:I|Q <voltage> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58

:INPut:OFFSet:I|Q? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58

:INPut[1]:IQ:BALanced[:STATe] OFF|ON|0|1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51

:INPut[1]:IQ:BALanced[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51

:INPut[1]:IQ:Q:IMPedance LOW|HIGH. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

:INPut[1]:IQ:Q:IMPedance? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

:INPut[1]:IQ:TYPE IQ|I|Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

:INPut[1]:IQ:TYPE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

:INPut[1]:IQ[:I]:IMPedance LOW|HIGH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51

:INPut[1]:IQ[:I]:IMPedance? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51

:INSTrument:CATalog? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 216

:INSTrument:COUPle:DEFault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229

:INSTrument:COUPle:FREQuency:CENTer ALL|NONE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228

:INSTrument:COUPle:FREQuency:CENTer?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228

:INSTrument:DEFault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135

:INSTrument:NSELect <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 216

:INSTrument:NSELect? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 216

:INSTrument[:SELect] SA|BASIC|WCDMA|CDMA2K|EDGEGSM|PNOISE|CDMA1XEV|CW-

LAN|WIMAXOFDMA|CWIMAXOFDM|VSA|VSA89601|LTE|IDEN|WIMAXFIXED|LTETDD|

TDSCDMA|NFIGURE|ADEMOD|DVB|DTMB|ISDBT|CMMB|RLC . . . . . . . . . . . . . . . . . . . 213

31

List of Commands

:INSTrument[:SELect]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .213

:LXI:CLOCk:PTP:ACCuracy

NS25|NS100|NS250|NS1000|NS2500|US10|US25|US100|US250|US1000|US2500|MS10|M

S25|MS100|MS1000|S10|GT10S|UNKNown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .238

:LXI:CLOCk:PTP:ACCuracy? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .238

:LXI:CLOCk:PTP:ANNounce:INTerval <interval> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .238

:LXI:CLOCk:PTP:ANNounce:INTerval?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .238

:LXI:CLOCk:PTP:ANNounce:RTOut <numberOfIntervals> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .239

:LXI:CLOCk:PTP:ANNounce:RTOut? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .239

:LXI:CLOCk:PTP:CCLass? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .239

:LXI:CLOCk:PTP:DEViation? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .239

:LXI:CLOCk:PTP:DOMain <domainNumber>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .240

:LXI:CLOCk:PTP:DOMain?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .240

:LXI:CLOCk:PTP:DRINterval <seconds>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .242

:LXI:CLOCk:PTP:DRINterval? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .243

:LXI:CLOCk:PTP:GMASter:ACCuracy? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .243

:LXI:CLOCk:PTP:GMASter:MADDress? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .243

:LXI:CLOCk:PTP:GMASter:TRACeability? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .244

:LXI:CLOCk:PTP:MASTer:MADDress? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .244

:LXI:CLOCk:PTP:OFFSet? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .240

:LXI:CLOCk:PTP:PRIority:FIRSt <priority> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .240

:LXI:CLOCk:PTP:PRIority:FIRSt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .240

:LXI:CLOCk:PTP:PRIority:SECond <priority> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .241

:LXI:CLOCk:PTP:PRIority:SECond? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .241

:LXI:CLOCk:PTP:SINTerval <seconds> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .242

:LXI:CLOCk:PTP:SINTerval? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .242

:LXI:CLOCk:PTP:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .241

:LXI:CLOCk:PTP:TRACeability? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .241

:LXI:CLOCk:PTP:VARiance?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .242

:LXI:CLOCk:SALGorithm:LOG:ALL? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .247

:LXI:CLOCk:SALGorithm:LOG:CIRCular:FBENtry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .245

:LXI:CLOCk:SALGorithm:LOG:CIRCular[:ENABled] ON|OFF|0|1 . . . . . . . . . . . . . . . . . . . . . . .245

32

List of Commands

:LXI:CLOCk:SALGorithm:LOG:CIRCular[:ENABled]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 245

:LXI:CLOCk:SALGorithm:LOG:CLEar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 245

:LXI:CLOCk:SALGorithm:LOG:COUNt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 246

:LXI:CLOCk:SALGorithm:LOG:ENABled ON|OFF|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 246

:LXI:CLOCk:SALGorithm:LOG:ENABled?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 246

:LXI:CLOCk:SALGorithm:LOG:ENTRy? <intIndex> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 247

:LXI:CLOCk:SALGorithm:LOG:SIZE <maxLogEntries>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 246

:LXI:CLOCk:SALGorithm:LOG:SIZE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 246

:LXI:CLOCk:SALGorithm:LOG:STATistics:CLEar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 247

:LXI:CLOCk:SALGorithm:LOG:STATistics[:DATA]?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 247

:LXI:CLOCk:SALGorithm:LOG[:NEXT]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 245

:LXI:CLOCk:SALGorithm[:SET]:ASYMmetry <seconds> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248

:LXI:CLOCk:SALGorithm[:SET]:ASYMmetry? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248

:LXI:CLOCk:SALGorithm[:SET]:CFTHreshold <secondsSquared> . . . . . . . . . . . . . . . . . . . . . . . . 248

:LXI:CLOCk:SALGorithm[:SET]:CFTHreshold? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248

:LXI:CLOCk:SALGorithm[:SET]:CIConstant <servoConstant> . . . . . . . . . . . . . . . . . . . . . . . . . . . 249

:LXI:CLOCk:SALGorithm[:SET]:CIConstant? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 249

:LXI:CLOCk:SALGorithm[:SET]:CONFigure <asymmetry>, <coarse fine threshold>, <cpc>, <cic>,

<fpc>, <fic>, <maximum outlier discard>, <outlier threshold>, <set/steer threshold> . . . . . . . . . 252

:LXI:CLOCk:SALGorithm[:SET]:CPConstant <servoConstant>. . . . . . . . . . . . . . . . . . . . . . . . . . . 249

:LXI:CLOCk:SALGorithm[:SET]:CPConstant?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 249

:LXI:CLOCk:SALGorithm[:SET]:FIConstant <servoConstant> . . . . . . . . . . . . . . . . . . . . . . . . . . . 250

:LXI:CLOCk:SALGorithm[:SET]:FIConstant? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 250

:LXI:CLOCk:SALGorithm[:SET]:FPConstant <servoConstant> . . . . . . . . . . . . . . . . . . . . . . . . . . . 250

:LXI:CLOCk:SALGorithm[:SET]:FPConstant?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 250

:LXI:CLOCk:SALGorithm[:SET]:OMAXimum <consecutiveSamples>. . . . . . . . . . . . . . . . . . . . . . 250

:LXI:CLOCk:SALGorithm[:SET]:OMAXimum? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 250

:LXI:CLOCk:SALGorithm[:SET]:OTENable ON|OFF|1|0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 251

:LXI:CLOCk:SALGorithm[:SET]:OTENable? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 251

:LXI:CLOCk:SALGorithm[:SET]:OTHReshold <standardDeviations>. . . . . . . . . . . . . . . . . . . . . . 251

:LXI:CLOCk:SALGorithm[:SET]:OTHReshold? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 251

33

List of Commands

:LXI:CLOCk:SALGorithm[:SET]:STHReshold <seconds>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .252

:LXI:CLOCk:SALGorithm[:SET]:STHReshold? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .252

:LXI:CLOCk:SYNC:LOCal:ENABled ON|OFF|0|1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .253

:LXI:CLOCk:SYNC:LOCal:ENABled? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .253

:LXI:CLOCk:SYNC:LOCal:INTerval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .253

:LXI:CLOCk:SYNC:LOCal:INTerval?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .253

:LXI:CLOCk:SYNC:MASTer? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .252

:LXI:CLOCk[:TIME]:DLSavings? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .235

:LXI:CLOCk[:TIME]:FRACtion <fraction> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .234

:LXI:CLOCk[:TIME]:FRACtion? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .234

:LXI:CLOCk[:TIME]:LOCal? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .234

:LXI:CLOCk[:TIME]:LSOFfset <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .234

:LXI:CLOCk[:TIME]:LSOFfset? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .234

:LXI:CLOCk[:TIME]:MARKer[1]|2|3|4|5|6|7|8|9:CLEAr . . . . . . . . . . . . . . . . . . . . . . . . . . . . .236

:LXI:CLOCk[:TIME]:MARKer[1]|2|3|4|5|6|7|8|9:DELTa? . . . . . . . . . . . . . . . . . . . . . . . . . . . .237

:LXI:CLOCk[:TIME]:MARKer[1]|2|3|4|5|6|7|8|9[:SET] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .236

:LXI:CLOCk[:TIME]:MEASure:CLEar. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .237

:LXI:CLOCk[:TIME]:MEASure[:DELTa]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .237

:LXI:CLOCk[:TIME]:SEConds <seconds> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .233

:LXI:CLOCk[:TIME]:SEConds?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .233

:LXI:CLOCk[:TIME]:TAI? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .235

:LXI:CLOCk[:TIME]:TZONe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .235

:LXI:CLOCk[:TIME]:UTC? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .235

:LXI:CLOCk[:TIME][:VALue] <seconds>,<fractionalSeconds>. . . . . . . . . . . . . . . . . . . . . . . . . . . . .232

:LXI:CLOCk[:TIME][:VALue]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .233

:LXI:EVENt:DOMain <intDomain> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .224

:LXI:EVENt:DOMain? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .224

:LXI:EVENt:INPut:LAN:ADD “LANEVENT” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .255

:LXI:EVENt:INPut:LAN:COUNt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .259

:LXI:EVENt:INPut:LAN:DISable:ALL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .255

:LXI:EVENt:INPut:LAN:FILTer “LANEVENT”,“filterString” . . . . . . . . . . . . . . . . . . . . . . . . . . . . .257

34

List of Commands

:LXI:EVENt:INPut:LAN:FILTer? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 257

:LXI:EVENt:INPut:LAN:IDENtifier “LANEVENT”,”identifier” . . . . . . . . . . . . . . . . . . . . . . . . . . . 257

:LXI:EVENt:INPut:LAN:IDENtifier? “LANEVENT” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 257

:LXI:EVENt:INPut:LAN:LIST? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 259

:LXI:EVENt:INPut:LAN:LOCation “LANEVENT”,”path” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 255

:LXI:EVENt:INPut:LAN:LOCation? “LANEVENT”. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 255

:LXI:EVENt:INPut:LAN:REMove:ALL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 256

:LXI:EVENt:INPut:LAN:REMove[:EVENt] “LANEVENT” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 256

:LXI:EVENt:INPut:LAN[:SET]:CONFigure “lanEvent” , <enab>, <detection>, <filter>, <identifier>

260

:LXI:EVENt:INPut:LAN[:SET]:DETection “LANEVENT”, HIGH|LOW|RISE|FALL . . . . . . . . 258

:LXI:EVENt:INPut:LAN[:SET]:DETection? “LANEVENT” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 258

:LXI:EVENt:INPut:LAN[:SET]:ENABled “LANEVENT”,ON|OFF|1|0 . . . . . . . . . . . . . . . . . . . . 259

:LXI:EVENt:INPut:LAN[:SET]:ENABled? “LANEVENT”. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 259

:LXI:EVENt:LOG:ALL? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183

:LXI:EVENt:LOG:CIRCular:FBENtry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184

:LXI:EVENt:LOG:CIRCular[:ENABle] ON|OFF|1|0. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182

:LXI:EVENt:LOG:CIRCular[:ENABle]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182

:LXI:EVENt:LOG:CLEar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182

:LXI:EVENt:LOG:COUNt?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183

:LXI:EVENt:LOG:ENABle ON|OFF|1|0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183

:LXI:EVENt:LOG:ENABle? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183

:LXI:EVENt:LOG:ENTRy? <intIndex> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184

:LXI:EVENt:LOG:SIZE <size>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182

:LXI:EVENt:LOG:SIZE?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182

:LXI:EVENt:LOG[:NEXT]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183

:LXI:EVENt:STATus[:ENABled] “STATUSEVENT”,ON|OFF|1|0. . . . . . . . . . . . . . . . . . . . . . . . 254

:LXI:EVENt:STATus[:ENABled]? “STATUSEVENT” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254

:LXI:EVENt[:OUTPut]:LAN:ADD “LANEVENT” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 226

:LXI:EVENt[:OUTPut]:LAN:COUNt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 231

:LXI:EVENt[:OUTPut]:LAN:DISable:ALL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225

35

List of Commands

:LXI:EVENt[:OUTPut]:LAN:LIST? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .225

:LXI:EVENt[:OUTPut]:LAN:REMove:ALL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .227

:LXI:EVENt[:OUTPut]:LAN:REMove[:EVENt] “LANEVENT” . . . . . . . . . . . . . . . . . . . . . . . . . . . .226

:LXI:EVENt[:OUTPut]:LAN:SEND “LANEVENT”", RISE|FALL . . . . . . . . . . . . . . . . . . . . . . . . . .232

:LXI:EVENt[:OUTPut]:LAN[:SET]:CONFigure “lanEvent”,<enabled>,<source>,<slope>,<drive>,<destination> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .231

:LXI:EVENt[:OUTPut]:LAN[:SET]:DESTination “LANEVENT”,”destinationExpression” . . . . . .228

:LXI:EVENt[:OUTPut]:LAN[:SET]:DESTination? “LANEVENT” . . . . . . . . . . . . . . . . . . . . . . . . . .228

:LXI:EVENt[:OUTPut]:LAN[:SET]:DRIVe “LANEVENT”, OFF|NORMal|WOR . . . . . . . . . . . . .229

:LXI:EVENt[:OUTPut]:LAN[:SET]:DRIVe? “LANEVENT”. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .229

:LXI:EVENt[:OUTPut]:LAN[:SET]:ENABled “LANEVENT”,ON|OFF|1|0. . . . . . . . . . . . . . . . . .231

:LXI:EVENt[:OUTPut]:LAN[:SET]:ENABled? “LANEVENT” . . . . . . . . . . . . . . . . . . . . . . . . . . . . .231

:LXI:EVENt[:OUTPut]:LAN[:SET]:IDENtifier “LANEVENT”, “identifier”. . . . . . . . . . . . . . . . . . .232

:LXI:EVENt[:OUTPut]:LAN[:SET]:IDENtifier? "LANEVENT" . . . . . . . . . . . . . . . . . . . . . . . . . . . .232

:LXI:EVENt[:OUTPut]:LAN[:SET]:SLOPe “LANEVENT”, POSitive|NEGative . . . . . . . . . . . . . .230

:LXI:EVENt[:OUTPut]:LAN[:SET]:SLOPe? “LANEVENT” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .230

:LXI:EVENt[:OUTPut]:LAN[:SET]:SOURce “LANEVENT”, “SourceEvent” . . . . . . . . . . . . . . . . . .228

:LXI:EVENt[:OUTPut]:LAN[:SET]:SOURce? “LANEVENT” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .228

:LXI:EVENt[:OUTPut]:LAN[:SET]:TSDelta “LANEVENT”, <seconds> . . . . . . . . . . . . . . . . . . . . .230

:LXI:EVENt[:OUTPut]:LAN[:SET]:TSDelta? “LANEVENT” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .230

:MEASure:W11A[n]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .445

:MEASure:W11B[n]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .397

:MEMMory:RDIRectory <directory_name> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .175

:MMEMory:CATalog? [<directory_name>] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .173

:MMEMory:CDIRectory [<directory_name>] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .173

:MMEMory:CDIRectory? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .173

:MMEMory:COPY <string>,<string>[,<string>,<string>]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .173

:MMEMory:DATA <file_name>, <data>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .174

:MMEMory:DATA? <file_name> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .174

:MMEMory:DELete <file_name>[,<directory_name>] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .174

:MMEMory:LOAD:CORRection 1|2|3|4, <filename> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .255

36

List of Commands

:MMEMory:LOAD:STATe <filename>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142

:MMEMory:LOAD:TRACe TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6,<filename>

145

:MMEMory:LOAD:TRACe:DATA D1|D2|D3|D4|D5|D6,<filename>[,CSV|TXT|SDF|MAT4|MAT|HDF5|BIN] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 411

:MMEMory:LOAD:TRACe:DATA TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6,<filename> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147

:MMEMory:LOAD:TRACe:REGister

TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6,<integer>. . . . . . . . . . . . . . . . . . . . . 145

:MMEMory:MDIRectory <directory_name> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175

:MMEMory:MOVE <string>,<string>[,<string>,<string>] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175

:MMEMory:STORe:CORRection 1|2|3|4, <filename> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 260

:MMEMory:STORe:SCReen <filename> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170

:MMEMory:STORe:SCReen:THEMe TDColor|TDMonochrome|FCOLor|FMONochrome . . . . . 168

:MMEMory:STORe:SCReen:THEMe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168

:MMEMory:STORe:STATe <filename> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158

:MMEMory:STORe:TRACe TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6|ALL,<filename> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161

:MMEMory:STORe:TRACe:DATA

TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6,"<filename>"[,CSV|TXT|SDF|MAT4

|MAT|HDF5|BIN[,OFF|ON|0|1]] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 414

:MMEMory:STORe:TRACe:DATA

TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6|ALL,<filename> . . . . . . . . . . . . . . 162

:MMEMory:STORe:TRACe:REGister

TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6|ALL,<integer> . . . . . . . . . . . . . . . . 161

:OUTPut:ANALog OFF|SVIDeo|LOGVideo|LINVideo|DAUDio . . . . . . . . . . . . . . . . . . . . . . . . . . 92

:OUTPut:ANALog:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93

:OUTPut:ANALog:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93

:OUTPut:ANALog? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92

:OUTPut:AUX SIF|AIF|LOGVideo|OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99

:OUTPut:AUX:AIF <value> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100

:OUTPut:AUX:AIF? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100

:OUTPut:AUX? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99

:OUTPut:DBUS[1][:STATe] ON|OFF|1|0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98

37

List of Commands

:OUTPut:DBUS[1][:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .98

:OUTPut:IQ:OUTPut IQ1|IQ250|OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97

:OUTPut:IQ:OUTPut? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97

:READ:<meas>[n]?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .482

:READ:ADEMod[n]?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .315

:READ:DDEMod[n]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .339

:READ:VECTor[n]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .287

:READ:W11A[n]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .445

:READ:W11B[n]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .397

:SOURce:NOISe:SNS:ATTached? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .212

:SOURce:NOISe:TYPE NORMal | SNS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .211

:SOURce:NOISe:TYPE?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .211

:SOURce:NOISe[:STATe] ON|OFF|1|0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .212

:SOURce:NOISe[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .212

:SYSTem:APPLication:CATalog:OPTion? <model> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .219

:SYSTem:APPLication:CATalog:REVision? <model>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .219

:SYSTem:APPLication:CATalog[:NAME]:COUNt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .218

:SYSTem:APPLication:CATalog[:NAME]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .218

:SYSTem:APPLication[:CURRent]:OPTion? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .218

:SYSTem:APPLication[:CURRent]:REVision? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .217

:SYSTem:APPLication[:CURRent][:NAME]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .217

:SYSTem:COMMunicate:GPIB[1][:SELF]:ADDRess <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . .218

:SYSTem:COMMunicate:GPIB[1][:SELF]:ADDRess? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .218

:SYSTem:COMMunicate:GPIB[1][:SELF]:CONTroller[:ENABle] ON|OFF|0|1 . . . . . . . . . . . . . .219

:SYSTem:COMMunicate:GPIB[1][:SELF]:CONTroller[:ENABle]?. . . . . . . . . . . . . . . . . . . . . . . . . .219

:SYSTem:COMMunicate:LAN:SCPI:SICL:ENABle OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . .221

:SYSTem:COMMunicate:LAN:SCPI:SICL:ENABle? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .221

:SYSTem:COMMunicate:LAN:SCPI:SOCKet:CONTrol?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .221

:SYSTem:COMMunicate:LAN:SCPI:SOCKet:ENABle OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . .220

:SYSTem:COMMunicate:LAN:SCPI:SOCKet:ENABle? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .220

:SYSTem:COMMunicate:LAN:SCPI:TELNet:ENABle OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . .220

38

List of Commands

:SYSTem:COMMunicate:LAN:SCPI:TELNet:ENABle? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 220

:SYSTem:COMMunicate:USB:CONNection? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222

:SYSTem:COMMunicate:USB:PACKets? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223

:SYSTem:COMMunicate:USB:STATus? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223

:SYSTem:DATE “<year>,<month>,<day>” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 274

:SYSTem:DATE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 274

:SYSTem:DEFault [ALL]|ALIGn|INPut|MISC|MODes|PON . . . . . . . . . . . . . . . . . . . . . . . . . . . 260

:SYSTem:ERRor:VERBose OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178

:SYSTem:ERRor:VERBose? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178

:SYSTem:ERRor[:NEXT]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177

:SYSTem:HELP:HEADers? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 273

:SYSTem:HID? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 267

:SYSTem:KLOCk OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 273

:SYSTem:KLOCk?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 273

:SYSTem:LKEY <”OptionInfo”>, <”LicenseInfo”> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 265

:SYSTem:LKEY:DELete <”OptionInfo”>,<”LicenseInfo”> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 266

:SYSTem:LKEY:LIST? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 266

:SYSTem:LKEY? <”OptionInfo”> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 267

:SYSTem:MRELay:COUNt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 270

:SYSTem:OPTions? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 272

:SYSTem:PON:APPLication:LLISt <string of INSTrument:SELect names> . . . . . . . . . . . . . . . . . 191

:SYSTem:PON:APPLication:LLISt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191

:SYSTem:PON:APPLication:VMEMory:TOTal? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191

:SYSTem:PON:APPLication:VMEMory:USED:NAME? <INSTrument:SELect name>. . . . . . . . . 192

:SYSTem:PON:APPLication:VMEMory:USED? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192

:SYSTem:PON:APPLication:VMEMory[:AVAilable]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191

:SYSTem:PON:ETIMe?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 271

:SYSTem:PON:MODE SA|BASIC|ADEMOD|NFIGURE|PNOISE|CDMA2K|TDSCD-

MA|VSA|VSA89601|WCDMA|WIMAXOFDMA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186

:SYSTem:PON:MODE?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186

:SYSTem:PON:TIME?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204

39

List of Commands

:SYSTem:PON:TYPE MODE|USER|LAST|PRESet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .184

:SYSTem:PON:TYPE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .184

:SYSTem:PRESet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .133

:SYSTem:PRESet:USER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .275

:SYSTem:PRESet:USER:ALL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .276

:SYSTem:PRESet:USER:SAVE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .277

:SYSTem:PRINt:THEMe TDColor|TDMonochrome|FCOLor|FMONochrome . . . . . . . . . . . . . . .131

:SYSTem:PRINt:THEMe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .131

:SYSTem:PUP:PROCess . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .189

:SYSTem:SECurity:USB:WPRotect[:ENABle] ON|OFF|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .268

:SYSTem:SECurity:USB:WPRotect[:ENABle]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .268

:SYSTem:SHOW OFF|ERRor|SYSTem|HARDware|LXI|HWSTatistics|ALIGNment|SOFTware

176

:SYSTem:SHOW? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .176

:SYSTem:TEMPerature:HEXTreme? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .271

:SYSTem:TEMPerature:LEXTreme? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .270

:SYSTem:TIME “<hour>,<minute>,<second>”. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .274

:SYSTem:TIME? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .274

:SYSTem:VERSion? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .273

:TRIGger:<meas>[:SEQuence]:SOURce IMMediate|VIDeo|EXTernal1 . . . . . . . . . . . . . . . . . . . . .471

:TRIGger:<meas>[:SEQuence]:SOURce? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .471

:TRIGger:<measurement>[:SEQuence]:IQ:SOURce EXTernal1|EXTernal2|IMMediate|IQMag|IDEMod|QDEMod|IINPut|QINPut|AIQMag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .296

:TRIGger:<measurement>[:SEQuence]:IQ:SOURce? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .296

:TRIGger:<measurement>[:SEQuence]:RF:SOURce EXTernal1|EXTernal2|IMMediate|LINE|FRAMe|RFBurst|VIDeo|IF|ALARm|LAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .294

:TRIGger:<measurement>[:SEQuence]:RF:SOURce? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .294

:TRIGger:<measurement>[:SEQuence]:SOURce EXTernal1|EXTernal2|IMMediate|LINE|FRAMe|RFBurst|VIDeo|IF|ALARm|LAN|IQMag|IDEMod|QDEMod|IINPut|QIN

Put|AIQMag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .290

:TRIGger:<measurement>[:SEQuence]:SOURce? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .290

:TRIGger[:SEQuence]:AIQMag:BANDwidth <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .345

:TRIGger[:SEQuence]:AIQMag:BANDwidth? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .345

40

List of Commands

:TRIGger[:SEQuence]:AIQMag:CENTer <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 345

:TRIGger[:SEQuence]:AIQMag:CENTer? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 345

:TRIGger[:SEQuence]:AIQMag:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 345

:TRIGger[:SEQuence]:AIQMag:DELay:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 345

:TRIGger[:SEQuence]:AIQMag:DELay:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 345

:TRIGger[:SEQuence]:AIQMag:DELay? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 345

:TRIGger[:SEQuence]:AIQMag:LEVel <ampl > . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 344

:TRIGger[:SEQuence]:AIQMag:LEVel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 344

:TRIGger[:SEQuence]:AIQMag:SLOPe POSitive | NEGative. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 344

:TRIGger[:SEQuence]:AIQMag:SLOPe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 344

:TRIGger[:SEQuence]:ATRigger <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 347

:TRIGger[:SEQuence]:ATRigger:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 347

:TRIGger[:SEQuence]:ATRigger:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 347

:TRIGger[:SEQuence]:ATRigger?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 347

:TRIGger[:SEQuence]:EXTernal1:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 305

:TRIGger[:SEQuence]:EXTernal1:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 477

:TRIGger[:SEQuence]:EXTernal1:DELay:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . 305

:TRIGger[:SEQuence]:EXTernal1:DELay:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . 477

:TRIGger[:SEQuence]:EXTernal1:DELay:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 305

:TRIGger[:SEQuence]:EXTernal1:DELay:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 477

:TRIGger[:SEQuence]:EXTernal1:DELay? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 305

:TRIGger[:SEQuence]:EXTernal1:DELay? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 477

:TRIGger[:SEQuence]:EXTernal1:HOLDoff <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 477

:TRIGger[:SEQuence]:EXTernal1:HOLDoff:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . 477

:TRIGger[:SEQuence]:EXTernal1:HOLDoff:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 477

:TRIGger[:SEQuence]:EXTernal1:HOLDoff:TYPE BELow|ABOVe . . . . . . . . . . . . . . . . . . . . . . . . 478

:TRIGger[:SEQuence]:EXTernal1:HOLDoff:TYPE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 478

:TRIGger[:SEQuence]:EXTernal1:HOLDoff? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 477

:TRIGger[:SEQuence]:EXTernal1:LEVel <level> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 304

:TRIGger[:SEQuence]:EXTernal1:LEVel <voltage> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 476

:TRIGger[:SEQuence]:EXTernal1:LEVel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 304

41

List of Commands

:TRIGger[:SEQuence]:EXTernal1:LEVel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .476

:TRIGger[:SEQuence]:EXTernal1:SLOPe POSitive|NEGative . . . . . . . . . . . . . . . . . . . . . . . . . . . .305

:TRIGger[:SEQuence]:EXTernal1:SLOPe POSitive|NEGative . . . . . . . . . . . . . . . . . . . . . . . . . . . .476

:TRIGger[:SEQuence]:EXTernal1:SLOPe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .305

:TRIGger[:SEQuence]:EXTernal1:SLOPe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .476

:TRIGger[:SEQuence]:EXTernal2:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .308

:TRIGger[:SEQuence]:EXTernal2:DELay:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . .308

:TRIGger[:SEQuence]:EXTernal2:DELay:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .308

:TRIGger[:SEQuence]:EXTernal2:DELay? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .308

:TRIGger[:SEQuence]:EXTernal2:LEVel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .307

:TRIGger[:SEQuence]:EXTernal2:LEVel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .307

:TRIGger[:SEQuence]:EXTernal2:SLOPe POSitive|NEGative . . . . . . . . . . . . . . . . . . . . . . . . . . . .307

:TRIGger[:SEQuence]:EXTernal2:SLOPe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .307

:TRIGger[:SEQuence]:FRAMe:ADJust <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .316

:TRIGger[:SEQuence]:FRAMe:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .320

:TRIGger[:SEQuence]:FRAMe:DELay:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .320

:TRIGger[:SEQuence]:FRAMe:DELay:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .320

:TRIGger[:SEQuence]:FRAMe:DELay? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .320

:TRIGger[:SEQuence]:FRAMe:OFFSet <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .315

:TRIGger[:SEQuence]:FRAMe:OFFSet:DISPlay:RESet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .317

:TRIGger[:SEQuence]:FRAMe:OFFSet? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .315

:TRIGger[:SEQuence]:FRAMe:PERiod <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .314

:TRIGger[:SEQuence]:FRAMe:PERiod? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .314

:TRIGger[:SEQuence]:FRAMe:SYNC EXTernal1|EXTernal2|RFBurst|OFF . . . . . . . . . . . . . . . .318

:TRIGger[:SEQuence]:FRAMe:SYNC:HOLDoff <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .320

:TRIGger[:SEQuence]:FRAMe:SYNC:HOLDoff:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . .320

:TRIGger[:SEQuence]:FRAMe:SYNC:HOLDoff:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .320

:TRIGger[:SEQuence]:FRAMe:SYNC:HOLDoff? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .320

:TRIGger[:SEQuence]:FRAMe:SYNC? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .318

:TRIGger[:SEQuence]:HOLDoff <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .347

:TRIGger[:SEQuence]:HOLDoff:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .347

42

List of Commands

:TRIGger[:SEQuence]:HOLDoff:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 347

:TRIGger[:SEQuence]:HOLDoff:TYPE NORMal|ABOVe|BELow . . . . . . . . . . . . . . . . . . . . . . . . . 348

:TRIGger[:SEQuence]:HOLDoff:TYPE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 348

:TRIGger[:SEQuence]:HOLDoff? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 347

:TRIGger[:SEQuence]:IDEMod:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 339

:TRIGger[:SEQuence]:IDEMod:DELay:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 339

:TRIGger[:SEQuence]:IDEMod:DELay:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 339

:TRIGger[:SEQuence]:IDEMod:DELay? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 339

:TRIGger[:SEQuence]:IDEMod:LEVel <voltage> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 338

:TRIGger[:SEQuence]:IDEMod:LEVel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 338

:TRIGger[:SEQuence]:IDEMod:SLOPe POSitive | NEGative . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 339

:TRIGger[:SEQuence]:IDEMod:SLOPe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 339

:TRIGger[:SEQuence]:IINPut:DELay <time>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 342

:TRIGger[:SEQuence]:IINPut:DELay:STATe OFF|ON|0|1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 342

:TRIGger[:SEQuence]:IINPut:DELay:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 342

:TRIGger[:SEQuence]:IINPut:DELay? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 342

:TRIGger[:SEQuence]:IINPut:LEVel <voltage> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 341

:TRIGger[:SEQuence]:IINPut:LEVel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 341

:TRIGger[:SEQuence]:IINPut:SLOPe POSitive | NEGative . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 341

:TRIGger[:SEQuence]:IINPut:SLOPe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 341

:TRIGger[:SEQuence]:IQMag:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 338

:TRIGger[:SEQuence]:IQMag:DELay:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 338

:TRIGger[:SEQuence]:IQMag:DELay:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 338

:TRIGger[:SEQuence]:IQMag:DELay? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 338

:TRIGger[:SEQuence]:IQMag:LEVel <ampl >. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 337

:TRIGger[:SEQuence]:IQMag:LEVel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 337

:TRIGger[:SEQuence]:IQMag:SLOPe POSitive | NEGative . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 337

:TRIGger[:SEQuence]:IQMag:SLOPe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 337

:TRIGger[:SEQuence]:LINE:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 303

:TRIGger[:SEQuence]:LINE:DELay:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 303

:TRIGger[:SEQuence]:LINE:DELay:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 303

43

List of Commands

:TRIGger[:SEQuence]:LINE:DELay? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303

:TRIGger[:SEQuence]:LINE:SLOPe POSitive|NEGative. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303

:TRIGger[:SEQuence]:LINE:SLOPe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303

:TRIGger[:SEQuence]:LXI:ALARm:DISable:ALL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .328

:TRIGger[:SEQuence]:LXI:ALARm:LIST? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .328

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:CONFigure "alarmEvent", <enable>, <epochSeconds>,

<epochFraction>, <period>, <repeat> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .336

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:ENABled "alarmEvent",ON|OFF|1|0. . . . . . . . . . . . .335

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:ENABled? "alarmEvent" . . . . . . . . . . . . . . . . . . . . . . . .335

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:PERiod "alarmEvent",<seconds>. . . . . . . . . . . . . . . . . .334

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:PERiod? "alarmEvent" . . . . . . . . . . . . . . . . . . . . . . . . . .334

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:REPeat "alarmEvent",<repetitions> . . . . . . . . . . . . . . .334

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:REPeat? "alarmEvent" . . . . . . . . . . . . . . . . . . . . . . . . . .335

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME:SEConds "alarmEvent",<seconds> . . . . . . . . . . .332

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME:SEConds? "alarmEvent" . . . . . . . . . . . . . . . . . . .332

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue] "alarmEvent",<seconds>, <fractionalSeconds>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .331

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]:ABSolute "alarmEvent","date","time" .329

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]:ABSolute? "alarmEvent" . . . . . . . . . . .329

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]:FRACtion "alarmEvent",<fractionalSeconds>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .333

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]:FRACtion? "alarmEvent" . . . . . . . . . . .333

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]:RELative "alarmEvent",<seconds> . . .333

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]:RELative? "alarmEvent" . . . . . . . . . . .334

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .332

:TRIGger[:SEQuence]:LXI:LAN:ADD "LANEVENT" . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .325

:TRIGger[:SEQuence]:LXI:LAN:COUNt?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .326

:TRIGger[:SEQuence]:LXI:LAN:DISable:ALL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .322

:TRIGger[:SEQuence]:LXI:LAN:LIST? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .322

:TRIGger[:SEQuence]:LXI:LAN:REMove:ALL. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .325

:TRIGger[:SEQuence]:LXI:LAN:REMove[:EVENt] "LANEVENT". . . . . . . . . . . . . . . . . . . . . . . . . .325

:TRIGger[:SEQuence]:LXI:LAN[:SET]:CONFigure "lanEvent", <enable>, <detection>, <delay>,<fil-

44

List of Commands

ter>,<identifier> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 327

:TRIGger[:SEQuence]:LXI:LAN[:SET]:DELay "LANEVENT",<time> . . . . . . . . . . . . . . . . . . . . . . 323

:TRIGger[:SEQuence]:LXI:LAN[:SET]:DELay? "lanEvent". . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 324

:TRIGger[:SEQuence]:LXI:LAN[:SET]:DETection "LANEVENT", HIGH|LOW|RISE|FALL. . . 323

:TRIGger[:SEQuence]:LXI:LAN[:SET]:DETection? "LANEVENT" . . . . . . . . . . . . . . . . . . . . . . . . . 323

:TRIGger[:SEQuence]:LXI:LAN[:SET]:ENABled "LANEVENT",ON|OFF|1|0 . . . . . . . . . . . . . . 324

:TRIGger[:SEQuence]:LXI:LAN[:SET]:ENABled? "LANEVENT" . . . . . . . . . . . . . . . . . . . . . . . . . . 324

:TRIGger[:SEQuence]:LXI:LAN[:SET]:FILTer "LANEVENT","filterString" . . . . . . . . . . . . . . . . . 326

:TRIGger[:SEQuence]:LXI:LAN[:SET]:FILTer? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 326

:TRIGger[:SEQuence]:LXI:LAN[:SET]:IDENtifier "LANEVENT","identifier" . . . . . . . . . . . . . . . . 327

:TRIGger[:SEQuence]:LXI:LAN[:SET]:IDENtifier? "LANEVENT". . . . . . . . . . . . . . . . . . . . . . . . . 327

:TRIGger[:SEQuence]:OFFSet <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 349

:TRIGger[:SEQuence]:OFFSet:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 349

:TRIGger[:SEQuence]:OFFSet:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 349

:TRIGger[:SEQuence]:OFFSet? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 349

:TRIGger[:SEQuence]:QDEMod:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 340

:TRIGger[:SEQuence]:QDEMod:DELay:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 340

:TRIGger[:SEQuence]:QDEMod:DELay:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 340

:TRIGger[:SEQuence]:QDEMod:DELay? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 340

:TRIGger[:SEQuence]:QDEMod:LEVel <voltage> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 340

:TRIGger[:SEQuence]:QDEMod:LEVel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 340

:TRIGger[:SEQuence]:QDEMod:SLOPe POSitive | NEGative . . . . . . . . . . . . . . . . . . . . . . . . . . . . 340

:TRIGger[:SEQuence]:QDEMod:SLOPe?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 340

:TRIGger[:SEQuence]:QINPut:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 343

:TRIGger[:SEQuence]:QINPut:DELay:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 343

:TRIGger[:SEQuence]:QINPut:DELay:STATe?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 343

:TRIGger[:SEQuence]:QINPut:DELay?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 343

:TRIGger[:SEQuence]:QINPut:LEVel <voltage> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 342

:TRIGger[:SEQuence]:QINPut:LEVel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 342

:TRIGger[:SEQuence]:QINPut:SLOPe POSitive | NEGative . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 343

:TRIGger[:SEQuence]:QINPut:SLOPe?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 343

45

List of Commands

:TRIGger[:SEQuence]:RFBurst:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .312

:TRIGger[:SEQuence]:RFBurst:DELay:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .312

:TRIGger[:SEQuence]:RFBurst:DELay:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .312

:TRIGger[:SEQuence]:RFBurst:DELay?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .312

:TRIGger[:SEQuence]:RFBurst:LEVel:ABSolute <ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .309

:TRIGger[:SEQuence]:RFBurst:LEVel:ABSolute? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .309

:TRIGger[:SEQuence]:RFBurst:LEVel:RELative <rel_ampl>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .311

:TRIGger[:SEQuence]:RFBurst:LEVel:RELative? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .311

:TRIGger[:SEQuence]:RFBurst:LEVel:TYPE ABSolute|RELative . . . . . . . . . . . . . . . . . . . . . . . . .310

:TRIGger[:SEQuence]:RFBurst:LEVel:TYPE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .310

:TRIGger[:SEQuence]:RFBurst:SLOPe POSitive|NEGative . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .312

:TRIGger[:SEQuence]:RFBurst:SLOPe?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .312

:TRIGger[:SEQuence]:RLINe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .471

:TRIGger[:SEQuence]:RLINe?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .471

:TRIGger[:SEQuence]:VIDeo:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .301

:TRIGger[:SEQuence]:VIDeo:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .473

:TRIGger[:SEQuence]:VIDeo:DELay:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .301

:TRIGger[:SEQuence]:VIDeo:DELay:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .473

:TRIGger[:SEQuence]:VIDeo:DELay:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .301

:TRIGger[:SEQuence]:VIDeo:DELay:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .473

:TRIGger[:SEQuence]:VIDeo:DELay?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .301

:TRIGger[:SEQuence]:VIDeo:DELay?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .473

:TRIGger[:SEQuence]:VIDeo:HOLDoff <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .474

:TRIGger[:SEQuence]:VIDeo:HOLDoff:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .474

:TRIGger[:SEQuence]:VIDeo:HOLDoff:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .474

:TRIGger[:SEQuence]:VIDeo:HOLDoff:TYPE BELow|ABOVe . . . . . . . . . . . . . . . . . . . . . . . . . . . .475

:TRIGger[:SEQuence]:VIDeo:HOLDoff:TYPE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .475

:TRIGger[:SEQuence]:VIDeo:HOLDoff? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .474

:TRIGger[:SEQuence]:VIDeo:LEVel <ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .299

:TRIGger[:SEQuence]:VIDeo:LEVel <voltage> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .472

:TRIGger[:SEQuence]:VIDeo:LEVel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .299

46

List of Commands

:TRIGger[:SEQuence]:VIDeo:LEVel?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 472

:TRIGger[:SEQuence]:VIDeo:SLOPe POSitive|NEGative . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300

:TRIGger[:SEQuence]:VIDeo:SLOPe POSitive|NEGative . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 473

:TRIGger[:SEQuence]:VIDeo:SLOPe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300

:TRIGger[:SEQuence]:VIDeo:SLOPe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 473

:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut HSWP|MEASuring|MAIN|GATE|GTRigger|OEVen|SSWeep|SSETtled|S1Marker|S2Marker|S3Marker|S4Marker|OFF. . . . . . . . . . . 90

:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut:POLarity POSitive|NEGative . . . . . . . . . . 90

:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut:POLarity? . . . . . . . . . . . . . . . . . . . . . . . . . . 90

:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90

:TRIGger1|TRIGger[:SEQuence]:LXI:ALARm:COUNt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 336

:UNIT:POWer DBM|DBMV|DBMA|V|W|A|DBUV|DBUA|DBUVM|DBUAM|DBPT|DBG . 21

:UNIT:POWer? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

[:SENSe]:<meas>:AVERage:COUNt <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 401

[:SENSe]:<meas>:AVERage:COUNt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 401

[:SENSe]:<meas>:AVERage:FAST OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 404

[:SENSe]:<meas>:AVERage:FAST:URATe <integer>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 404

[:SENSe]:<meas>:AVERage:FAST:URATe:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . 404

[:SENSe]:<meas>:AVERage:FAST:URATe:AUTO?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 404

[:SENSe]:<meas>:AVERage:FAST:URATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 404

[:SENSe]:<meas>:AVERage:FAST? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 404

[:SENSe]:<meas>:AVERage:TCONtrol EXPonential|REPeat. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 402

[:SENSe]:<meas>:AVERage:TCONtrol? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 402

[:SENSe]:<meas>:AVERage:TYPE RMS|TIME|MAXimum. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 403

[:SENSe]:<meas>:AVERage:TYPE?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 403

[:SENSe]:<meas>:AVERage[:STATe] OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 401

[:SENSe]:<meas>:AVERage[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 401

[:SENSe]:<meas>:BANDwidth|BWIDth[:RESolution] <bandwidth>. . . . . . . . . . . . . . . . . . . . . . . 369

[:SENSe]:<meas>:BANDwidth|BWIDth[:RESolution]:COUPle SPAN|MIN|FIXed . . . . . . . . . . 370

[:SENSe]:<meas>:BANDwidth|BWIDth[:RESolution]:COUPle? . . . . . . . . . . . . . . . . . . . . . . . . . . 370

[:SENSe]:<meas>:BANDwidth|BWIDth[:RESolution]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 369

47

List of Commands

[:SENSe]:<meas>:FFT:WINDow[:TYPE] UNIForm|HANNing|GAUSsian|FLATtop. . . . . . . . . .371

[:SENSe]:<meas>:FFT:WINDow[:TYPE]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .371

[:SENSe]:<meas>:PUPDate:ENABle OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .499

[:SENSe]:<meas>:PUPDate:ENABle? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .499

[:SENSe]:<meas>:SWEep:EGATe:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .427

[:SENSe]:<meas>:SWEep:EGATe:DELay? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .427

[:SENSe]:<meas>:SWEep:EGATe:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .426

[:SENSe]:<meas>:SWEep:EGATe:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .426

[:SENSe]:<meas>:SWEep:EGATe[:SPAN] <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .426

[:SENSe]:<meas>:SWEep:EGATe[:SPAN]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .426

[:SENSe]:<meas>:SWEep:ISRate? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .499

[:SENSe]:<meas>:SWEep:POINts <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .427

[:SENSe]:<meas>:SWEep:POINts:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .427

[:SENSe]:<meas>:SWEep:POINts:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .427

[:SENSe]:<meas>:SWEep:POINts?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .427

[:SENSe]:<meas>:SWEep:TIME <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .425

[:SENSe]:<meas>:SWEep:TIME? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .425

[:SENSe]:ADC:DITHer:AUTO[:STATe] OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .200

[:SENSe]:ADC:DITHer:AUTO[:STATe]?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .200

[:SENSe]:ADC:DITHer[:STATe] OFF|ON|HIGH. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .199

[:SENSe]:ADC:DITHer[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .199

[:SENSe]:ADEMod:AM:UNIT AM|PCT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .327

[:SENSe]:ADEMod:AM:UNIT? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .327

[:SENSe]:ADEMod:CARRier:FREQuency:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . .328

[:SENSe]:ADEMod:CARRier:FREQuency:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .328

[:SENSe]:ADEMod:CARRier:PHASe:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .329

[:SENSe]:ADEMod:CARRier:PHASe:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .329

[:SENSe]:ADEMod:MODulation AM|FM|PM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .326

[:SENSe]:ADEMod:MODulation? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .326

[:SENSe]:AVERage:CLEar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .169

[:SENSe]:AVERage:COUNt <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .168

48

List of Commands

[:SENSe]:AVERage:COUNt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168

[:SENSe]:AVERage:TYPE RMS|LOG|SCALar [:SENSe]:AVERage:TYPE? . . . . . . . . . . . . . . . . . 170

[:SENSe]:AVERage:TYPE:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170

[:SENSe]:AVERage:TYPE:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170

[:SENSe]:BANDwidth|BWIDth:[:RESolution]:MODE EMI|SAN|OFF . . . . . . . . . . . . . . . . . . . . 237

[:SENSe]:BANDwidth|BWIDth:[:RESolution]:MODE?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237

[:SENSe]:CORRection:BTS[:RF]:GAIN <rel_ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65

[:SENSe]:CORRection:BTS[:RF]:GAIN? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65

[:SENSe]:CORRection:CSET:ALL:DELete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

[:SENSe]:CORRection:CSET:ALL[:STATe] ON|OFF|1|0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82

[:SENSe]:CORRection:CSET:ALL[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82

[:SENSe]:CORRection:CSET[1]|2|3|4:ANTenna[:UNIT] GAUSs|PTESla|UVM|UAM|NOConversion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

[:SENSe]:CORRection:CSET[1]|2|3|4:ANTenna[:UNIT]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:COMMent "text" . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:COMMent? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DATA <freq>, <ampl>, . . . . . . . . . . . . . . . . . . . . . . . . 83

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DATA:MERGe <freq>, <ampl>, ... . . . . . . . . . . . . . . . 84

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DATA? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DELete. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DESCription "text" . . . . . . . . . . . . . . . . . . . . . . . . . . . 79

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DESCription? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:X:SPACing LINear|LOGarithmic . . . . . . . . . . . . . . . 77

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:X:SPACing? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6[:STATe] ON|OFF|1|0 . . . . . . . . . . . . . . . . . . . . . . . . 74

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74

[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude] 50|75 . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

[:SENSe]:CORRection:IQ:I:GAIN <rel_ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66

[:SENSe]:CORRection:IQ:I:GAIN? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66

[:SENSe]:CORRection:IQ:I|Q:ATTenuation <rel_ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

49

List of Commands

[:SENSe]:CORRection:IQ:I|Q:ATTenuation:RATio <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57

[:SENSe]:CORRection:IQ:I|Q:ATTenuation:RATio? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57

[:SENSe]:CORRection:IQ:I|Q:ATTenuation?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57

[:SENSe]:CORRection:IQ:Q:GAIN <rel_ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67

[:SENSe]:CORRection:IQ:Q:GAIN? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67

[:SENSe]:CORRection:IQ:Q:SKEW <seconds> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56

[:SENSe]:CORRection:IQ:Q:SKEW? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56

[:SENSe]:CORRection:IQ[:I]:SKEW <seconds> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52

[:SENSe]:CORRection:IQ[:I]:SKEW? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52

[:SENSe]:CORRection:MS[:RF]:GAIN <rel_ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64

[:SENSe]:CORRection:MS[:RF]:GAIN? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64

[:SENSe]:CORRection:NOISe:FLOor ON|OFF|1|0. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .240

[:SENSe]:CORRection:NOISe:FLOor? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .240

[:SENSe]:CORRection:SA[:RF]:GAIN <rel_ampl>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63

[:SENSe]:CORRection:SA[:RF]:GAIN? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63

[:SENSe]:DDEMod:ALPHa <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .362

[:SENSe]:DDEMod:ALPHa? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .362

[:SENSe]:DDEMod:APSK:R2Ratio <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .369

[:SENSe]:DDEMod:APSK:R2Ratio? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .369

[:SENSe]:DDEMod:APSK:R3Ratio <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .369

[:SENSe]:DDEMod:APSK:R3Ratio? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .369

[:SENSe]:DDEMod:CADJust <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .367

[:SENSe]:DDEMod:CADJust? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .367

[:SENSe]:DDEMod:EQUalization:CONVergence <real>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .373

[:SENSe]:DDEMod:EQUalization:CONVergence? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .373

[:SENSe]:DDEMod:EQUalization:FLENgth <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .373

[:SENSe]:DDEMod:EQUalization:FLENgth?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .373

[:SENSe]:DDEMod:EQUalization:HOLD OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .374

[:SENSe]:DDEMod:EQUalization:HOLD? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .374

[:SENSe]:DDEMod:EQUalization:RESet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .374

[:SENSe]:DDEMod:EQUalization:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .372

50

List of Commands

[:SENSe]:DDEMod:EQUalization:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 372

[:SENSe]:DDEMod:FILTer:MEASurement NONE|RRCosine|GAUSsian|EDGE|IS95EQ|RECTangle|LPASs|USER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 359

[:SENSe]:DDEMod:FILTer:MEASurement:REGister D1|D2|D3|D4|D5|D6 . . . . . . . . . . . . . . . 360

[:SENSe]:DDEMod:FILTer:MEASurement:REGister?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 360

[:SENSe]:DDEMod:FILTer:MEASurement? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 359

[:SENSe]:DDEMod:FILTer:REFerence RCOSine|RRCosine|GAUSsian|EDGE|IS95BB|RECTangle|HSINe|USER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 361

[:SENSe]:DDEMod:FILTer:REFerence:REGister D1|D2|D3|D4|D5|D6. . . . . . . . . . . . . . . . . . . 362

[:SENSe]:DDEMod:FILTer:REFerence:REGister? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 362

[:SENSe]:DDEMod:FILTer:REFerence? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 361

[:SENSe]:DDEMod:LSNR OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 370

[:SENSe]:DDEMod:LSNR? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 370

[:SENSe]:DDEMod:MODulation

BPSK|QPSK|OQPSK|PI4DQPSK|DQPSK|PSK8|PI8DPSK8|DPSK8|QAM16|QAM32|QAM6

4|QAM128|QAM256|QAM512|QAM1024|FSK2|FSK4|FSK8|FSK16|MSK1|MSK2|EDGE|A

PSK16|APSK32|DVBAPSK16|DVBAPSK32|DVBQAM16|DVBQAM32|DVBQAM64|DVBQAM

128|DVBQAM256|VSB8|VSB16 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 354

[:SENSe]:DDEMod:MODulation? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 354

[:SENSe]:DDEMod:PPSYmbol <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 357

[:SENSe]:DDEMod:PPSYmbol? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 357

[:SENSe]:DDEMod:SRATe <frequency> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 355

[:SENSe]:DDEMod:SRATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 355

[:SENSe]:DDEMod:STANdard:PRESet CDMABTS|CDMAMS|CD-

PD|EDGE|GSM|NADC|PDC|PHS|WCDMA|BLUETOOTH|HIPERLANHBR|HIPERLANLBR

|WLAN11B|ZIGBEE2450|ZIGBEE868|ZIGBEE915|DTV8|DTV16|DVB16|DVB32|DVB64|D

VB128|DVB256|DVB16APSK23|DVB16APSK34|DVB16APSK45|DVB16APSK56|DVB16APSK

89|DVB16APSK910|DVB32APSK34|DVB32APSK45|DVB32APSK56|DVB32APSK89|DVB32A

PSK910|APCO|DECT|TETRA|VDL3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 375

[:SENSe]:DDEMod:SWEep:POINts <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 356

[:SENSe]:DDEMod:SWEep:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 356

[:SENSe]:DDEMod:SYNC:BURSt:STATe OFF|ON|0|1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 363

[:SENSe]:DDEMod:SYNC:BURSt:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 363

[:SENSe]:DDEMod:SYNC:SLENgth <time>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 363

[:SENSe]:DDEMod:SYNC:SLENgth? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 363

51

List of Commands

[:SENSe]:DDEMod:SYNC:SWORd:OFFSet <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .366

[:SENSe]:DDEMod:SYNC:SWORd:OFFSet? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .366

[:SENSe]:DDEMod:SYNC:SWORd:PATTern <string> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .365

[:SENSe]:DDEMod:SYNC:SWORd:PATTern? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .365

[:SENSe]:DDEMod:SYNC:SWORd:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .364

[:SENSe]:DDEMod:SYNC:SWORd:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .364

[:SENSe]:DEMod AM|FM|PM|OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .205

[:SENSe]:DEMod:AM:BANDwidth:CHANnel <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .206

[:SENSe]:DEMod:AM:BANDwidth:CHANnel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .206

[:SENSe]:DEMod:FM:BANDwidth:CHANnel <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .207

[:SENSe]:DEMod:FM:BANDwidth:CHANnel?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .207

[:SENSe]:DEMod:FM:DEEMphasis OFF|US25|US50|US75|US750 . . . . . . . . . . . . . . . . . . . . . .207

[:SENSe]:DEMod:FM:DEEMphasis? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .207

[:SENSe]:DEMod:PM:BANDwidth:CHANnel <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .209

[:SENSe]:DEMod:PM:BANDwidth:CHANnel?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .209

[:SENSe]:DEMod:STATe OFF|ON|0|1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .210

[:SENSe]:DEMod:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .210

[:SENSe]:DEMod:TIME <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .210

[:SENSe]:DEMod:TIME? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .210

[:SENSe]:DEMod? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .205

[:SENSe]:EMC:STANdard[:SELect] NONE|CISPr|MIL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .236

[:SENSe]:EMC:STANdard[:SELect]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .236

[:SENSe]:FCAPture:BLOCk <integer>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .497

[:SENSe]:FCAPture:BLOCk? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .497

[:SENSe]:FCAPture:LENGth <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .496

[:SENSe]:FCAPture:LENGth? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .496

[:SENSe]:FCAPture:POINter <integer>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .497

[:SENSe]:FCAPture:POINter? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .497

[:SENSe]:FCAPture:WLENgth AUTO|BIT32|BIT64 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .496

[:SENSe]:FCAPture:WLENgth? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .496

[:SENSe]:FEED RF|AIQ|IQ|IONLy|QONLy|INDependent|AREFerence . . . . . . . . . . . . . . . . . .43

52

List of Commands

[:SENSe]:FEED:AREFerence REF50|REF4800|OFF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

[:SENSe]:FEED:AREFerence? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

[:SENSe]:FEED:DATA INPut|STORed|RECorded. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

[:SENSe]:FEED:DATA:STORe. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70

[:SENSe]:FEED:DATA? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

[:SENSe]:FEED:IQ:TYPE IQ|IONLy|QONLy|INDependent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

[:SENSe]:FEED:IQ:TYPE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

[:SENSe]:FEED? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

[:SENSe]:FREQuency:CENTer <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

[:SENSe]:FREQuency:CENTer <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 372

[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 374

[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

[:SENSe]:FREQuency:CENTer:STEP:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 374

[:SENSe]:FREQuency:CENTer:STEP:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 374

[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 374

[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

[:SENSe]:FREQuency:CENTer?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

[:SENSe]:FREQuency:CENTer?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 372

[:SENSe]:FREQuency:CISPr:BAND A|B|C|CD|D|E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237

[:SENSe]:FREQuency:IQ:CENTer <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

[:SENSe]:FREQuency:IQ:CENTer? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

[:SENSe]:FREQuency:OFFSet <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

[:SENSe]:FREQuency:OFFSet? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

[:SENSe]:FREQuency:RF:CENTer <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

[:SENSe]:FREQuency:RF:CENTer?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

[:SENSe]:FREQuency:SPAN <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 416

[:SENSe]:FREQuency:SPAN:FULL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 417

[:SENSe]:FREQuency:SPAN? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 416

[:SENSe]:FREQuency:STARt <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 373

53

List of Commands

[:SENSe]:FREQuency:STARt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .373

[:SENSe]:FREQuency:STOP <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .373

[:SENSe]:FREQuency:STOP? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .373

[:SENSe]:FREQuency:SYNThesis:AUTO[:STATe] OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . .197

[:SENSe]:FREQuency:SYNThesis:AUTO[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .197

[:SENSe]:FREQuency:SYNThesis[:STATe] 1|2|3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .196

[:SENSe]:FREQuency:SYNThesis[:STATe]?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .196

[:SENSe]:FREQuency:TUNE:IMMediate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36

[:SENSe]:IF:GAIN:FFT:AUTO[:STATe] OFF|ON|0|1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .204

[:SENSe]:IF:GAIN:FFT:AUTO[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .204

[:SENSe]:IF:GAIN:FFT[:STATe] AUTOrange|LOW|HIGH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .203

[:SENSe]:IF:GAIN:FFT[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .203

[:SENSe]:IF:GAIN:SWEPt:AUTO[:STATe] OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .202

[:SENSe]:IF:GAIN:SWEPt:AUTO[:STATe]?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .202

[:SENSe]:IF:GAIN:SWEPt[:STATe] OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .201

[:SENSe]:IF:GAIN:SWEPt[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .201

[:SENSe]:POWer:IQ:Q:RANGe[:UPPer] <ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15

[:SENSe]:POWer:IQ:Q:RANGe[:UPPer]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15

[:SENSe]:POWer:IQ:RANGe:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12

[:SENSe]:POWer:IQ:RANGe:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12

[:SENSe]:POWer:IQ[:I]:RANGe[:UPPer] <ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13

[:SENSe]:POWer:IQ[:I]:RANGe[:UPPer]?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13

[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4

[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4

[:SENSe]:POWer[:RF]:ATTenuation:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4

[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement] 10 dB | 2 dB. . . . . . . . . . . . . . . . . . . . . . .10

[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10

[:SENSe]:POWer[:RF]:ATTenuation? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4

[:SENSe]:POWer[:RF]:EATTenuation <rel_ampl> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8

[:SENSe]:POWer[:RF]:EATTenuation:STATe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6

[:SENSe]:POWer[:RF]:EATTenuation:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6

54

List of Commands

[:SENSe]:POWer[:RF]:EATTenuation? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

[:SENSe]:POWer[:RF]:GAIN:BAND LOW|FULL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

[:SENSe]:POWer[:RF]:GAIN:BAND?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

[:SENSe]:POWer[:RF]:GAIN[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer] <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

[:SENSe]:POWer[:RF]:MW:PATH STD|LNPath|MPBypass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

[:SENSe]:POWer[:RF]:MW:PATH?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

[:SENSe]:POWer[:RF]:PADJust <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

[:SENSe]:POWer[:RF]:PADJust:PRESelector MWAVe|MMWave|EXTernal. . . . . . . . . . . . . . . . . . 20

[:SENSe]:POWer[:RF]:PADJust:PRESelector? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

[:SENSe]:POWer[:RF]:PADJust? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

[:SENSe]:POWer[:RF]:PCENter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

[:SENSe]:POWer[:RF]:RANGe <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 361

[:SENSe]:POWer[:RF]:RANGe:AUTO ON|OFF|1|0. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

[:SENSe]:POWer[:RF]:RANGe:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

[:SENSe]:POWer[:RF]:RANGe:OPTimize IMMediate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

[:SENSe]:POWer[:RF]:RANGe:OPTimize:ATTenuation OFF|ELECtrical|COMBined . . . . . . . . . . 9

[:SENSe]:POWer[:RF]:RANGe:OPTimize:ATTenuation? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

[:SENSe]:POWer[:RF]:RANGe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 361

[:SENSe]:RADio:STANdard:BAND:CLASs BC0|BC1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232

[:SENSe]:RADio:STANdard:BAND:CLASs? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232

[:SENSe]:RADio:STANdard:DEVice BTS|MS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 231

[:SENSe]:RADio:STANdard:DEVice? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 231

[:SENSe]:RADio:STANdard:EAMeas YES|NO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236

[:SENSe]:RADio:STANdard:EAMeas?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236

[:SENSe]:RADio:STANdard:PACKet DH1|DH3|DH5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234

[:SENSe]:RADio:STANdard:PACKet? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234

[:SENSe]:RADio:STANdard[:SELect]

NONE|JSTD|IS95a|IS97D|IS98D|GSM|W3GPP|CDMA2000MC1|C20001X|NADC|PDC|BL

UEtooth|TETRa|WL802DOT11A|WL802DOT11B|WL802DOT11G|HIPERLAN2|DVBTLSN|D

55

List of Commands

VBTGPN|DVBTIPN|FCC15|SDMBSE|UWBINDOOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .230

[:SENSe]:RADio:STANdard[:SELect]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .230

[:SENSe]:RECording:ABORt. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71

[:SENSe]:RECording:INITiate[:IMMediate] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71

[:SENSe]:RECording:LENGth <real>,SEConds|RECords|POINts . . . . . . . . . . . . . . . . . . . . . . . . .72

[:SENSe]:RECording:LENGth:STATe MAX|MANual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72

[:SENSe]:RECording:LENGth:STATe?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72

[:SENSe]:RECording:LENGth:UNIT? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72

[:SENSe]:RECording:LENGth:VALue?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72

[:SENSe]:ROSCillator:BANDwidth WIDE|NARRow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87

[:SENSe]:ROSCillator:BANDwidth?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87

[:SENSe]:ROSCillator:COUPling NORMal|NACQuisition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .89

[:SENSe]:ROSCillator:COUPling? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .89

[:SENSe]:ROSCillator:EXTernal:FREQuency <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87

[:SENSe]:ROSCillator:EXTernal:FREQuency? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87

[:SENSe]:ROSCillator:SOURce INTernal|EXTernal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85

[:SENSe]:ROSCillator:SOURce:TYPE INTernal|EXTernal|SENSe . . . . . . . . . . . . . . . . . . . . . . . . .85

[:SENSe]:ROSCillator:SOURce:TYPE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85

[:SENSe]:ROSCillator:SOURce? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85

[:SENSe]:SWEep:EGATe:CONTrol EDGE|LEVel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .284

[:SENSe]:SWEep:EGATe:CONTrol? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .284

[:SENSe]:SWEep:EGATe:DELay <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .280

[:SENSe]:SWEep:EGATe:DELay:COMPensation:TYPE OFF|SETTled|GDELay . . . . . . . . . . . . .286

[:SENSe]:SWEep:EGATe:DELay:COMPensation:TYPE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .286

[:SENSe]:SWEep:EGATe:DELay? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .280

[:SENSe]:SWEep:EGATe:HOLDoff <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .285

[:SENSe]:SWEep:EGATe:HOLDoff:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .285

[:SENSe]:SWEep:EGATe:HOLDoff:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .285

[:SENSe]:SWEep:EGATe:HOLDoff? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .285

[:SENSe]:SWEep:EGATe:LENGth <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .281

[:SENSe]:SWEep:EGATe:LENGth? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .281

56

List of Commands

[:SENSe]:SWEep:EGATe:METHod LO|VIDeo|FFT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281

[:SENSe]:SWEep:EGATe:METHod? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281

[:SENSe]:SWEep:EGATe:MINFast? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 288

[:SENSe]:SWEep:EGATe:SOURce EXTernal1|EXTernal2 |LINE|FRAMe|RFBurst . . . . . . . . . 283

[:SENSe]:SWEep:EGATe:SOURce? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 283

[:SENSe]:SWEep:EGATe:TIME <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 280

[:SENSe]:SWEep:EGATe:TIME? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 280

[:SENSe]:SWEep:EGATe:VIEW ON|OFF|1|0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 277

[:SENSe]:SWEep:EGATe:VIEW? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 277

[:SENSe]:SWEep:EGATe[:STATe] OFF|ON|0|1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 276

[:SENSe]:SWEep:EGATe[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 276

[:SENSe]:SWEep:FFT:WIDTh <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 272

[:SENSe]:SWEep:FFT:WIDTh:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 273

[:SENSe]:SWEep:FFT:WIDTh:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 273

[:SENSe]:SWEep:FFT:WIDTh? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 272

[:SENSe]:SWEep:POINts <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 288

[:SENSe]:SWEep:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 288

[:SENSe]:SWEep:TIME <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264

[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264

[:SENSe]:SWEep:TIME:AUTO:RULes NORMal|ACCuracy|SRESponse . . . . . . . . . . . . . . . . . . . 266

[:SENSe]:SWEep:TIME:AUTO:RULes:AUTO[:STATe] ON|OFF|1|0 . . . . . . . . . . . . . . . . . . . . . . 267

[:SENSe]:SWEep:TIME:AUTO:RULes:AUTO[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 267

[:SENSe]:SWEep:TIME:AUTO:RULes? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 266

[:SENSe]:SWEep:TIME:AUTO?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264

[:SENSe]:SWEep:TIME? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264

[:SENSe]:SWEep:TYPE FFT|SWEep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 268

[:SENSe]:SWEep:TYPE:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 269

[:SENSe]:SWEep:TYPE:AUTO:RULes SPEed|DRANge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 270

[:SENSe]:SWEep:TYPE:AUTO:RULes:AUTO[:STATe] OFF|ON|0|1. . . . . . . . . . . . . . . . . . . . . . 271

[:SENSe]:SWEep:TYPE:AUTO:RULes:AUTO[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 271

[:SENSe]:SWEep:TYPE:AUTO:RULes? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 270

57

List of Commands

[:SENSe]:SWEep:TYPE:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .269

[:SENSe]:SWEep:TYPE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .268

[:SENSe]:VECTor|ADEMod:FREQuency:CENTer:TRACk OFF|ON|0|1 . . . . . . . . . . . . . . . . . . .417

[:SENSe]:VECTor|ADEMod:FREQuency:CENTer:TRACk?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .417

[:SENSe]:VOLTage:IQ:Q:RANGe[:UPPer] <voltage> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15

[:SENSe]:VOLTage:IQ:Q:RANGe[:UPPer]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15

[:SENSe]:VOLTage:IQ:RANGe:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12

[:SENSe]:VOLTage:IQ:RANGe:AUTO?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12

[:SENSe]:VOLTage:IQ[:I]:RANGe[:UPPer] <voltage> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13

[:SENSe]:VOLTage:IQ[:I]:RANGe[:UPPer]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13

[:SENSe]:VOLTage|POWer:IQ:MIRRored OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14

[:SENSe]:VOLTage|POWer:IQ:MIRRored? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14

[:SENSe]:W11A:EQUalizer:TRAining CHANnel|CDATa. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .471

[:SENSe]:W11A:EQUalizer:TRAining?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .471

[:SENSe]:W11A:GINTerval <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .460

[:SENSe]:W11A:GINTerval? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .460

[:SENSe]:W11A:STANdard I11AGOFDM | HIPERLAN2 | I11GDSSSOFDM | I11AGTURBO |

I11PDSRC | I11J10MHZ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .473

[:SENSe]:W11A:STANdard:PRESet I11AGOFDM|HIPERLAN2|

I11GDSSSOFDM|I11AGTURBO|I11PDSRC|I11J10MHZ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .461

[:SENSe]:W11A:STANdard? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .473

[:SENSe]:W11A:SUBCarrier:NUMBer <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .467

[:SENSe]:W11A:SUBCarrier:NUMBer? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .467

[:SENSe]:W11A:SUBCarrier:SELect ALL|PILot|SINGle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .466

[:SENSe]:W11A:SUBCarrier:SELect? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .466

[:SENSe]:W11A:SUBCarrier:SPACing <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .468

[:SENSe]:W11A:SUBCarrier:SPACing? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .468

[:SENSe]:W11A:SYNC:SEQuence LONG|SHORt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .472

[:SENSe]:W11A:SYNC:SEQuence? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .472

[:SENSe]:W11A:SYNC:SLENgth <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .463

[:SENSe]:W11A:SYNC:SLENgth?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .463

[:SENSe]:W11A:TADJust <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .469

58

List of Commands

[:SENSe]:W11A:TADJust? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 469

[:SENSe]:W11A:TIME:INTerval <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 466

[:SENSe]:W11A:TIME:INTerval? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 466

[:SENSe]:W11A:TIME:OFFSet <integer>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 465

[:SENSe]:W11A:TIME:OFFSet? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 465

[:SENSe]:W11A:TIME:RESult:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 464

[:SENSe]:W11A:TIME:RESult:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 464

[:SENSe]:W11A:TIME:RESult:LENGth <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 464

[:SENSe]:W11A:TIME:RESult:LENGth? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 464

[:SENSe]:W11A:TIME:RESult:MAX <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 465

[:SENSe]:W11A:TIME:RESult:MAX ?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 465

[:SENSe]:W11A:TRACk:AMPLitude OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 470

[:SENSe]:W11A:TRACk:AMPLitude? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 470

[:SENSe]:W11A:TRACk:PHASe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 470

[:SENSe]:W11A:TRACk:PHASe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 470

[:SENSe]:W11A:TRACk:TIMing OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 471

[:SENSe]:W11A:TRACk:TIMing?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 471

[:SENSe]:W11A[:SUBCarrier]:DEMod BPSK|QPSK|QAM16|QAM64 . . . . . . . . . . . . . . . . . . . . 459

[:SENSe]:W11A[:SUBCarrier]:DEMod:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 459

[:SENSe]:W11A[:SUBCarrier]:DEMod:AUTO?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 459

[:SENSe]:W11A[:SUBCarrier]:DEMod? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 459

[:SENSe]:W11B:ALPHa <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 415

[:SENSe]:W11B:ALPHa? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 415

[:SENSe]:W11B:CADJust <real> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 420

[:SENSe]:W11B:CADJust? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 420

[:SENSe]:W11B:CRATe <freq> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 419

[:SENSe]:W11B:CRATe?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 419

[:SENSe]:W11B:DEMod DSSS1M|DSSS2M|CCK5M5|CCK11M|PBCC5M5|

PBCC11M|PBCC22M|PBCC33M . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 411

[:SENSe]:W11B:DEMod:AUTO OFF|ON|0|1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 411

[:SENSe]:W11B:DEMod:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 411

59

List of Commands

[:SENSe]:W11B:DEMod? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .411

[:SENSe]:W11B:DESCramble ALL|NONE|PREamble|PHEader . . . . . . . . . . . . . . . . . . . . . . . . .421

[:SENSe]:W11B:DESCramble? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .421

[:SENSe]:W11B:EQUalization:FLENgth <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .421

[:SENSe]:W11B:EQUalization:FLENgth? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .421

[:SENSe]:W11B:EQUalization[:STATe] OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .421

[:SENSe]:W11B:EQUalization[:STATe]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .421

[:SENSe]:W11B:FILTer:REFerence RECTangular|GAUSsian|RCOSine . . . . . . . . . . . . . . . . . . . .414

[:SENSe]:W11B:FILTer:REFerence?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .414

[:SENSe]:W11B:STANdard:PRESet I11BGDSSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .415

[:SENSe]:W11B:SYNC:SLENgth <time> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .416

[:SENSe]:W11B:SYNC:SLENgth?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .416

[:SENSe]:W11B:TIME:INTerval <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .418

[:SENSe]:W11B:TIME:INTerval? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .418

[:SENSe]:W11B:TIME:OFFSet <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .418

[:SENSe]:W11B:TIME:OFFSet? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .418

[:SENSe]:W11B:TIME:RESult:AUTO OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .417

[:SENSe]:W11B:TIME:RESult:AUTO? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .417

[:SENSe]:W11B:TIME:RESult:LENGth <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .417

[:SENSe]:W11B:TIME:RESult:LENGth? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .417

[:SENSe]:W11B:TIME:RESult:MAX <integer> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .417

[:SENSe]:W11B:TIME:RESult:MAX ? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .417

[:SENSe]:W11B:TRACk:PHASe OFF|ON|0|1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .420

[:SENSe]:W11B:TRACk:PHASe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .420

FCAP:BLOC <num_points_per_read_block> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .495

FCAP:LENG <num_samples>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .495

FCAP:LENG? MAX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .496

FCAP:POIN <read_pointer_position> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .495

FCAP:POIN? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .495

FETC:FCAP? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .495

INIT:FCAP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .495

60

List of Commands

VECT:SWE:ISR? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 495

61

List of Commands

62

1

Using Help

Welcome to the X-Series Signal Analyzer Help system!

The online Help system is "context-sensitive". This means that the information displayed when you invoke the Help system depends on the selected Analyzer Mode, Measurement and key.

TIP

To view help for any front-panel key or menu key, press that key with this Help

Window open.

To scroll any page vertically (to see the whole of a long topic), press the

Down

Arrow

key on the front panel to scroll down (or the

Up Arrow

key to scroll up). To locate these keys, see

“Front Panel Keys used by the Help System” on page 71

.

See “Navigating the Help Window Without a Mouse” on page 76

for complete information about Using

Help without an attached Mouse and Keyboard. For specific details of how to navigate to topics, see

“Finding a Topic without a Mouse and Keyboard” on page 85 .

See “Navigating the Help Window with a Mouse” on page 74

to learn about Using Help with an

attached Mouse and Keyboard.

You can view Help on the Analyzer itself, or you can View Help on Another Computer, by copying the

Help files and viewing Help there. For details, see the Section

“Viewing Help on a separate Computer” on page 65 .

To locate Other Available Help Resources, see

“Locating Other Help Resources” on page 64

.

Key Path

Help

63

Using Help

Locating Other Help Resources

Locating Other Help Resources

All available documentation is present on the Analyzer’s hard disk, either as HTML Help or Acrobat

PDF files.

In addition to the interactive Windows (HTML) Help system, the Analyzer’s hard disk contains

Application Notes, tutorial documents, etc.

This same documentation is also included on the Documentation CD shipped with your Analyzer.

Many of the supporting documents use the Adobe Acrobat (PDF) file format. You can view PDF files using the pre-installed Adobe Reader software.

The Adobe Reader user interface differs from the Windows Help interface. For full details on how to

navigate within Acrobat documents using Adobe Reader, see “Navigating Acrobat (PDF) Files” on page 80 .

64 Chapter 1

Using Help

Viewing Help on a separate Computer

Viewing Help on a separate Computer

You may want to view the help pages without having them appear on top of the Analyzer’s screen.

There are two separate Help files for each Analyzer Mode, which contain all the same help pages in different formats:

1. A file in HTML Help (CHM) format,

2. A file in Acrobat (PDF) format.

You can copy any of the Help files to another computer, then open and view the help pages in the file on that computer.

Your choice of which file to copy and view may depend on what you want to do with the file (for example, whether you want to print it and read the paper copy, or view it on the computer). The table below compares the relative advantages of the two formats:

Format Type

File Extension

Software Required to view file

Full Text Search?

Printable?

Printable Table of

Contents?

Navigable without a

Mouse and

Keyboard?

HTML Help Format (CHM Files)

CHM

Microsoft Windows operating system only, with Microsoft Internet Explorer installed.

Yes

Yes, but with limited control.

No

Yes, but with some loss of functionality.

Has Page Numbers?

No

Context-Sensitive

Display?

Yes, when viewed using the X-Series

Analyzer application window.

Indexed?

Active Hyperlinks?

Yes

Yes

Acrobat Format (PDF Files)

PDF

Free Adobe Reader software can be downloaded for many operating systems, including: Microsoft Windows, Macintosh,

Linux, Solaris.

Yes

Yes. Full print control.

Yes

No

Yes

No

No

Yes

Copying the HTML Help (CHM) Files

You can copy the HTML Help file(s) you need to a separate computer running Microsoft Windows. Each

HTML Help file has a .chm extension.

You can find the HTML Help (.chm) files:

Either, on the documentation CD that came with the Analyzer,

Chapter 1 65

Using Help

Viewing Help on a separate Computer

Or, in a special directory on the Analyzer’s hard disk. The directory path is:

C:\Program Files\Agilent\SignalAnalysis\Infrastructure\Help

The illustration below shows an example listing of the HTML Help files in this directory, viewed using Windows Explorer.

Depending on which Analyzer software licenses you purchased, the content of the directory on your machine may vary.

NOTE

You can open and view the HTML Help files only on a PC that has Microsoft

Windows and Microsoft Internet Explorer installed.

Copying the Acrobat (PDF) Files

You can copy the Acrobat file(s) you need to a separate computer running any of several different operating systems. Each Acrobat file has a .pdf extension.

You can find the Acrobat (.pdf) files:

Either, on the documentation CD that came with the Analyzer,

Or, in a special directory on the Analyzer’s hard disk. The directory path is:

C:\Program Files\Agilent\SignalAnalysis\Infrastructure\Help\bookfiles

— The illustration below shows an example listing of the Acrobat files in this directory, viewed using Windows Explorer.

— The PDF versions of the help files are named <mode>_ref.pdf, where <mode> is the name of the Analyzer Mode. For example, the name of the PDF file for GSM/EDGE Mode is gsmedge_ref.pdf

. (Note that the directory also contains other PDF documents.)

— When you open any <mode>_ref.pdf document, the title page displays "<Mode> User’s and Programmer’s Reference

", where <Mode> is the name of the Analyzer Mode described by the document.

— Depending on which Analyzer software licenses you purchased, the content of the directory on your machine may vary.

66 Chapter 1

Using Help

Viewing Help on a separate Computer

Chapter 1 67

Using Help

How Help is Organized

How Help is Organized

This topic contains the following sections:

“Help Contents Listing” on page 68

“System Functions” on page 68

“Key Descriptions for Each Measurement” on page 69

“Key Information for Softkeys” on page 69

“Common Measurement Functions” on page 70

Help Contents Listing

The listing under the Contents tab in the Help Window includes a topic for each Front-panel key and each softkey, for each available measurement.

The Contents listing is split into several major sections, as shown below for the HTML Help version of the document. The structure of the PDF version is similar.

Help information is split between these sections as follows:

1. Using Help: this section.

2. System Functions. See “System Functions” on page 68

below.

3. Measurement Functions. See “Key Descriptions for Each Measurement” on page 69 below.

4. Common Measurement Functions. See “Common Measurement Functions” on page 70

below.

System Functions

This section contains information for the following keys, which are listed in alphabetical order:

File

,

68 Chapter 1

Using Help

How Help is Organized

Preset

,

Print

,

Quick Save

,

Recall

,

Save

,

System

,

User Preset

.

The functions of these keys do not vary between measurements: they operate the same way, irrespective of which Analyzer measurement you have selected.

The sections for

Recall

and

Save

contain only cross-references to the respective sections in

“Common

Measurement Functions” on page 70

, and are included here for convenience.

Key Descriptions for Each Measurement

The Contents section for each Measurement is sub-divided into topics for each Front-panel key, in alphabetical order, as shown below.

If you don’t see a topic for a Front-panel key in the Measurement-specific section, then it is located in the

section “System Functions” on page 68

.

Key Information for Softkeys

Information for each softkey that appears when you press a Front-panel key (or a softkey with a submenu) is listed under the entry for that key in the Help Contents. The example below shows the submenu under the

SPAN X Scale

Front-panel key in the "Waveform" Measurement, alongside the

Chapter 1 69

Using Help

How Help is Organized

actual softkeys for that menu.

In these subsections, all softkeys are listed in the order they appear in their menu (that is, not in alphabetical order).

Common Measurement Functions

This section groups together function and key information that is shared between measurements.

However, there is a listing for every Front-panel key and subkey in the section for each measurement, so you will generally not need to refer to this section.

The key subsections are listed alphabetically.

70 Chapter 1

Using Help

Front Panel Keys used by the Help System

Front Panel Keys used by the Help System

The interactive Help system uses the Front-panel keys shown below.

Item

Description

# Name

1 Help Key

2 Cancel (Esc)

Key

3 Next Window

Key

4 Arrow / Enter

Keys

5 Backward

Tab Key

6 Knob

7 Forward

Tab Key

8 Select /

Space Key

9 Ctrl Key

10 Alt Key

Opens Help (displaying the topic for the last key pressed).

Exits Help.

Changes the current window pane selection.

A central

Enter

key, surrounded by four directional arrow keys. Navigates within the Help system.

Moves between controls in the Help display.

For future use.

Moves between controls in the Help display.

Navigates within the Help system, in conjunction with other keys.

Navigates within the Help system, in conjunction with other keys. See

“Navigating Windows

HTML Help (CHM) Files” on page 72

.

Navigates within the Help system, in conjunction with other keys. See

“Navigating Windows

HTML Help (CHM) Files” on page 72

.

Acts as a "Back" key when navigating the pages of the Help system.

11 Bk Sp

(Backspace)

Key

Chapter 1 71

Using Help

Navigating Windows HTML Help (CHM) Files

Navigating Windows HTML Help (CHM) Files

HTML Help Window Components

When the interactive Help Window is open, the Analyzer’s display appears as below.

The HTML Help Window appears on top of, and to the left of, the measurement display. You can still see and use the current softkey menu when the HTML Help Window is open. However, pressing a softkey when the Help window is open displays Help for that softkey, but does not execute the softkey’s function.

When the Help Window is open, the Analyzer retains its current Mode and Measurement, as shown in the Title Bar.

The HTML Help Window itself consists of two panes, as shown in the diagram above.

On the left is the Navigation Pane, and on the right is the Topic Pane.

72 Chapter 1

Using Help

Navigating Windows HTML Help (CHM) Files

The Help Window Navigation Pane

The Navigation Pane is further divided into four tabs: Contents, Index, Search and Favorites, as shown below.

For details of how to switch between these tabs, if you don’t have a mouse attached to the Analyzer, see

the Section “To Switch the Active Tab within the Navigation Pane” on page 76

.

The Help Window Topic Pane

This pane displays the text for the topic that you have selected. It also contains clickable

Previous Page

and

Next Page

buttons (as shown below), which can be used to move to the previous or next page in the

Help file.

Basic Help Window Operations

This topic contains the following sections:

“Opening Help” on page 73

“Getting Help for a Specific Key” on page 73

“Closing the Help Window” on page 74

“Viewing Help on How to Use Help” on page 74

“Exiting Help on How to Use Help” on page 74

To locate the keys mentioned in this section, see

“Front Panel Keys used by the Help System” on page 71

.

Opening Help

To access the Help system, press the green

Help

key below the front panel display (shown below) while an Agilent application is running.

Note that the softkeys remain visible when the Help window is open.

Getting Help for a Specific Key

1. If the Help window is already open, press the desired key. The relevant Help topic appears.

Note that the function normally invoked by the key is not executed when the key is pressed with the

Help window open. If you want to execute the key’s function, first close Help by pressing the

Cancel

(Esc)

key (as described in “Closing the Help Window” on page 74

), then press the key, before opening Help again (if required).

2. If the Help window is not already open, press the desired key (which executes the key’s function), then press the

Help

key to display the relevant Help page. Help is available for all softkeys, and for

Chapter 1 73

Using Help

Navigating Windows HTML Help (CHM) Files

all the Front-panel keys listed under the "System Functions" and "Measurement" sections.

For details of how to navigate within the panes of the Help window, see “Navigating Windows HTML

Help (CHM) Files” on page 72 .

Closing the Help Window

To close the Help window, and return to the measurement application, press the

Cancel (Esc)

key

(depicted below).

Viewing Help on How to Use Help

With the Help window open, press the green

Help

key again.

The "Using Help" page appears, as shown below.

Exiting Help on How to Use Help

See the Section

“To Go Back or Forward: display the Previously-viewed or Next-viewed Topic in the

Topic Pane” on page 78

for details of several methods to accomplish this.

Navigating the Help Window

The way you navigate around the HTML Help Window depends on whether you have a mouse and keyboard attached to your Analyzer:

• If you have a mouse and keyboard attached, see the Section

“Navigating the Help Window with a

Mouse” on page 74

.

If you don’t have a mouse and keyboard attached, see the Section “Navigating the Help Window

Without a Mouse” on page 76

.

Navigating the Help Window with a Mouse

When the HTML Help window is open, you can point-and-click to navigate, as you would when using

Help for any Microsoft Windows computer application. The basic navigational features the Help systems of all X-Series Analyzers are as follows:

• If necessary, press the green

Help

key on the Front Panel, as described in

“Opening Help” on page 73

, to open the HTML Help window.

74 Chapter 1

Using Help

Navigating Windows HTML Help (CHM) Files

• Choose the desired topic from the list under the Contents Tab of the HTML Help Window’s

Navigation Pane, then click on the topic title to display the first page of the topic.

• To expand the listing of a topic, click on the + icon to the left of the topic’s book icon, as shown below. A list of subtopics and pages appears.

• To move to the Next or Previous Page within the Topic Pane, click the

Next Page

or

Previous Page

Keys (at the top right of the Topic Pane), as shown below.

Searching for a Help Topic If you also have a keyboard attached to the Analyzer, you can use the Help

system’s full-text search feature to locate help for any topic, by typing in a key name, a topic name, or any other desired text.

Select the "Search" tab of the Help window’s Navigation Pane, then use the following procedure:

1. Type the desired topic name into the Search window as shown in the diagram above. Note that the text search is not case-sensitive.

2. Click on the

List Topics

button.

3. Either:

Double-click on the desired topic in the list,

Chapter 1 75

Using Help

Navigating Windows HTML Help (CHM) Files

Or:

Click on the desired topic to select it, then click the

Display

button beneath the list.

4. The topic is then displayed in the Topic Pane (right-hand side of display).

Navigating the Help Window Without a Mouse

Most features of the Help system can be accessed and navigated without the necessity to attach a mouse or keyboard to the Analyzer. There are, however, a few exceptions to this rule, which are noted in the

Section

“Functions that cannot be used without a Mouse and Keyboard” on page 79

.

For information about how to perform common tasks in the Help system, click on one of the following links:

“To Toggle the Focus between the Navigation Pane and the Topic Pane” on page 76

“To Switch the Active Tab within the Navigation Pane” on page 76

“To Scroll up or down the list of Topics within the Contents or Index Tabs of the Navigation Pane” on page 77

“To Expand or Collapse a selected topic within the Contents Tab of the Navigation Pane” on page 77

“To Display a selected Help topic in the Topic Pane from the Contents Tab of the Navigation Pane” on page 77

“To Display a Help topic in the Topic Pane from the Index Tab of the Navigation Pane” on page 77

“To Scroll up or down within a topic in the Topic Pane” on page 77

“To Go to the Next or Previous Page in the Topic Pane” on page 78

“To Go Back or Forward: display the Previously-viewed or Next-viewed Topic in the Topic Pane” on page 78

“To Scroll horizontally or vertically within the Contents Tab of the Navigation Pane” on page 79

“To Print the topic currently displayed” on page 79

To locate all the keys mentioned in this section, see

“Front Panel Keys used by the Help System” on page 71 .

To Toggle the Focus between the Navigation Pane and the Topic Pane Press the

Next Window

key.

To Switch the Active Tab within the Navigation Pane Perform this procedure to display either the

Contents, Index, Search or Favorites tab of the Help window’s Navigation Pane.

Hold down the

Ctrl

key, then press either the

Forward Tab

key, or the

Backward Tab

key.

76 Chapter 1

Using Help

Navigating Windows HTML Help (CHM) Files

To Scroll up or down the list of Topics within the Contents or Index Tabs of the Navigation Pane

With the focus in the Navigation Pane, press the

Up Arrow

or

Down Arrow

keys.

To Expand or Collapse a selected topic within the Contents Tab of the Navigation Pane With the

focus in the Navigation Pane, press the

Right Arrow

key to expand the selected topic:

Or press the

Left Arrow

key to collapse the selected topic.

To Display a selected Help topic in the Topic Pane from the Contents Tab of the Navigation Pane

With the focus in the Contents Tab of the Navigation Pane, press the

Enter

key. If the selected topic was not already expanded, it expands in the Navigation Pane.

To Display a Help topic in the Topic Pane from the Index Tab of the Navigation Pane With the

focus in the Index Tab of the Navigation Pane, press the

Enter

key.

To Scroll up or down within a topic in the Topic Pane With the focus in the Topic Pane, press either

Chapter 1 77

Using Help

Navigating Windows HTML Help (CHM) Files

the

Up Arrow

key or

Down Arrow

key.

To Go to the Next or Previous Page in the Topic Pane With the focus in the Topic Pane, press either

Forward Tab

or

Backward Tab

keys to select the

>

(

Next Page

) key at the top right of the Pane, if you want to go to the next page, or select the

<

(

Previous Page

) key at the top right of the Pane, if you want to go to the previous page.

Press

Enter

.

To Go Back or Forward: display the Previously-viewed or Next-viewed Topic in the Topic Pane

To go back, either:

Hold down the

Alt

key, then press the

Left Arrow

key.

Or:

Press the

Bk Sp

key.

To go forward, hold down the

Alt

key, then press the

Right Arrow

key.

78 Chapter 1

Using Help

Navigating Windows HTML Help (CHM) Files

(The "Go Forward" operation has no effect unless there have been previous “Go Back” operations)

To Scroll horizontally or vertically within the Contents Tab of the Navigation Pane To scroll

horizontally: with the focus in the Contents Tab of the Navigation Pane, hold down the

Ctrl

key, then press either the

Left Arrow

or

Right Arrow

keys.

To scroll vertically: with the focus in the Contents Tab of the Navigation Pane, hold down the

Ctrl

key, then press either the

Up Arrow

or

Down Arrow

keys.

To Print the topic currently displayed Press the Front-panel

Print

key

Functions that cannot be used without a Mouse and Keyboard The following parts of the HTML

Help System cannot easily be used without attaching a mouse and keyboard to the Analyzer.

• The menu options at the top of the Help Window, consisting of:

Hide

,

Back

,

Print

and

Options

.

• The functionality of the Search Tab of the Navigation Pane.

• The functionality of the Favorites Tab of the Navigation Pane.

Chapter 1 79

Using Help

Navigating Acrobat (PDF) Files

Navigating Acrobat (PDF) Files

IMPORTANT

To navigate PDF files effectively, you must attach a mouse and keyboard to the

Analyzer.

If it is not possible to attach a mouse and keyboard to the Analyzer, you should copy the PDF file to a separate computer, then open it on that computer. Every

PDF file that is present on the Analyzer’s hard disk can also be found on the

Documentation CD shipped with the Analyzer. For details, see “Copying the

Acrobat (PDF) Files” on page 66

.

Adobe Reader Window

When an Adobe Acrobat (PDF) file is open and being viewed, the Analyzer’s display appears as below.

Note that, unlike the HTML Help Window, the Acrobat Reader Window is not embedded in the

Analyzer’s Application window. It is a separate window, which can be resized, moved and closed independently of the Application window.

The Adobe Reader Window itself consists of two panes, as shown in the diagram above.

On the left is the Navigation Pane (which may be hidden), and on the right is the Document Pane.

The Navigation Pane is further subdivided into four tabs: Bookmarks, Pages, Attachments and

Comments. Typically, PDF files supplied with the Agilent X-Series Analyzers contain useful content only under the Bookmarks and Pages Tabs: the Attachments and Comments Tabs are not used.

80 Chapter 1

Using Help

Navigating Acrobat (PDF) Files

Navigating the Acrobat Reader Window

The online Help for Adobe Reader provides detailed information on how to use the Reader. To access the online Help, do the following:

• With the Adobe Reader window open, click

Help

,

Adobe Reader Help

in the menu at the top of the screen. This opens the Help window on top of the document window.

• To close the Help window, either click the Red

X

at the top right of the window, or right-click anywhere in the title bar, then select

Close

from the popup menu.

Printing Acrobat Files

NOTE

The driver for the appropriate printer must be installed on the Analyzer’s hard disk before any file can be printed.

To print all or part of an open Acrobat file, do the following.

1. Either, a. click on the Print icon in the Acrobat Reader toolbar, b. or, select File > Print from the menu.

Chapter 1 81

Using Help

Navigating Acrobat (PDF) Files

2. The Acrobat Reader Print dialog opens, as shown below.

3. Choose the desired options within the Print dialog, then click OK to print (or click Cancel to cancel the printing).

NOTE

Clicking the Properties button within the Print dialog opens a window containing controls that are specific to the printer model installed. Check the printer manufacturer’s documentation for details of these capabilities.

82 Chapter 1

Using Help

Terms Used in This Documentation

Terms Used in This Documentation

Many special terms are used throughout this documentation. Please refer to the "Getting Started Guide" for detailed explanations of all these terms.

The Section below provides a brief description of special terms used in the Key parameter tables.

Terms used in Key Parameter Tables

The following terms are used in the parameter tables for each front-panel key or softkey. However, a particular key description may not use all the terms listed.

Term

Default Unit

Default

Terminator

Meaning

The default measurement unit of the setting.

Indicates the units that will be attached to the numeric value that you have entered. This default will be used from the front panel, when you terminate your entry by pressing the

Enter

key, rather then selecting a units key. This default will be used remotely when you send the command without specifying any units after your value(s).

Some commands may be unavailable when other parameters are set in certain ways. If applicable, any such limitations are described here.

Dependencies/

Couplings

Example

Factory Preset

Key Path

Knob Increment/Decrement

Max

Min

Meas Global

Meas Local

Mode Global

Preset

Range

Remote

Command

Provides command examples using the indicated remote command syntax.

Describes the function settings after a

Factory Preset

.

The sequence of Front-panel keys that accesses the function or setting.

The numeric value of the minimum increment or decrement that is applied when turning the thumb wheel knob.

The Maximum numerical value that the setting can take.

The Minimum numerical value that the setting can take.

The functionality described is the same in all measurements.

The functionality described is only true for the measurement selected.

The functionality described is the same for all modes.

In some cases, a Preset operation changes the status of a parameter. If the operation of the key specified is modified by a Preset operation, the effect is described here.

Describes the range of the smallest to largest values to which the function can be set. If you try to set a value below the minimum value, the analyzer defaults to the minimum value. If you try to set a value above the maximum value, the analyzer defaults to the maximum value.

Shows the syntax requirements for each SCPI command.

Chapter 1 83

Using Help

Terms Used in This Documentation

Term

Remote Command Notes

Resolution

SCPI Status Bits/OPC

Dependencies

State Saved

Meaning

Additional notes regarding Remote Commands.

Specifies the smallest change that can be made to the numeric value of a parameter.

Pressing certain keys may affect one or more status bits. If applicable, details are given here.

Indicates what happens to a particular function when the Analyzer state is saved (either to an external memory device or the internal D: drive). It also indicates whether the current settings of the function are maintained if the

Analyzer is powered on or preset using

Power On Last State

or

User

Preset

.

84 Chapter 1

Using Help

Context Sensitive Help not Available

Context Sensitive Help not Available

You have been directed to this page because interactive help for the key you selected is not available.

The following information may help you to find related topics of interest:

• If your Analyzer has an attached Mouse and Keyboard, see the Section

“Searching for a Help Topic” on page 75

.

If your Analyzer does not have an attached Mouse and Keyboard, see the Section “Finding a Topic without a Mouse and Keyboard” on page 85 below.

• If you want to learn how to select on-page links without a Mouse attached to your Analyzer, see the

Section

“Selecting a Hyperlink without a Mouse” on page 86

below.

TIP

If you want to understand the organization of Help, see the Section

“How Help is

Organized” on page 68 .

Finding a Topic without a Mouse and Keyboard

Follow this procedure when you want to display a different Help topic by selecting it from the Contents tab of the Help window’s Navigation Pane, but you do not have a mouse attached to the Analyzer.

Perform this action:

1. If necessary, toggle the focus between the Contents tab of the Navigation Pane (left side of display) and the Topic Pane

(right side of display) by pressing the

Next Window

key.

Ensure that the focus is in the Contents tab of the

Navigation Pane.

2. Move up or down the Contents list, by pressing the

Up

Arrow

or

Down Arrow

keys. Topics become highlighted upon selection.

Using these keys:

3. Display the selected topic, by pressing the

Enter

key.

Chapter 1 85

Using Help

Context Sensitive Help not Available

Selecting a Hyperlink without a Mouse

Follow this procedure when you want to select and follow a hyperlink on a Help page, but you do not have a mouse attached to the Analyzer.

Perform this action: Using these keys:

1. If necessary, toggle the focus between the Contents tab of the Navigation Pane (left side of display) and the Topic Pane

(right side of display) by pressing the

Next Window

key.

Ensure that the focus is in the Topic Pane.

2. Move from link to link in the Topic Pane (right side of display) by pressing the

Forward Tab

and

Backward Tab

keys. Links become highlighted upon selection.

NOTE: When a Help page is first displayed, no link is selected. Clicking the

Forward Tab

key once selects the

Previous Page

key. Clicking the

Forward Tab

key a second time selects the

Next Page

key. Clicking the

Forward Tab

key for a third time selects the first hyperlink on the page.

It is sometimes difficult to see the highlighting of the

Previous

and

Next Page

keys.

3. When you have selected the desired link, activate it by pressing the

Enter

key.

Use the

Forward

and

Backward Tab

keys to select the

Previous

and

Next Page

keys

86 Chapter 1

2

Additional Documentation

This section lists all available documentation for the X-Series Signal Analyzers.

All the documents listed in this section are installed on this Analyzer’s hard disk, and may be opened by clicking on one of the links. All the documents are also included on the Documentation CD shipped with your Analyzer.

For details of the types of documentation available, see

“Documentation Categories” on page 88 .

NOTE

If you do not have a mouse attached to your Analyzer, you may be unable to navigate or close PDF and CHM documents opened by the links on these pages. In this case, it is recommended that you should copy the PDF or CHM file to another computer that has an attached mouse.

The available documentation is divided into 3 groups as listed below. Click on a link for a detailed list of the documents in each group.

For All Measurement Applications

For Specific Measurement Applications

Application Notes & Other Documentation

87

Additional Documentation

Documentation Categories

Documentation Categories

All available documentation for X Series Analyzers falls into one of the following categories.

Type

Getting Started Guide

Specifications Guide

Functional Testing Guide

Instrument Messages

Measurement Guide

Programmer's Guide

User's and Programmer's

Reference

Description

Turn on process, Windows XP use/configuration, Front and Rear panel.

Specifications for all available Measurement Applications and optional hardware (for example, Spectrum Analyzer and W-CDMA).

Quick checks to verify overall instrument operation.

Descriptions of displayed messages of Information, Warnings and Errors.

Includes examples of measurements made using the front panel keys, or over a remote interface.

Provides information about how to program Agilent X-Series Signal

Analyzers with SCPI, and explains how to use the programming documentation. Describes programming examples, which are available in several languages.

Descriptions of front panel key functionality and the corresponding SCPI commands. Also includes some concept information.

88 Chapter 2

Additional Documentation

For All Measurement Applications

For All Measurement Applications

Click on any of the links below to open the document in a new window.

• Getting Started Guide [PDF]

• EXA Specifications Guide [PDF]

• MXA Specifications Guide [PDF]

• Instrument Messages Guide [PDF]

• CXA Functional Tests Guide [PDF]

• EXA Functional Tests Guide [PDF]

• MXA Functional Tests Guide [PDF]

• Programmer’s Guide [PDF]

• Programming Examples [Windows Explorer Folder]

• Programming Conversion Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88 .

Back to “Additional Documentation” on page 87 .

Chapter 2 89

Additional Documentation

For Specific Measurement Applications

For Specific Measurement Applications

For a list of links to all documents for a specific Measurement Application, click on one of the links below.

N6149A-2FP iDEN/WiDEN/MotoTalk Measurement Application

N6153A-2FP DVB-T/H Measurement Application

N6155A-2FP ISDB-T Measurement Application

N6156A-2FP DTMB Measurement Application

N6158A-2FP CMMB Measurement Application

N9020A/N9010A IQ Analyzer Measurement Application

N9020A/N9010A Spectrum Analyzer Measurement Application

N9061A Remote Language Compatibility Application

N9063A Analog Demod Measurement Application

N9068A Phase Noise Measurement Application

N9069A Noise Figure Measurement Application

N9071A GSM-EDGE Measurement Application

N9072A cdma2000 Measurement Application

N9073A W-CDMA Measurement Application

N9074A-XFP Single Acquisition Combined Fixed WiMAX Measurement Application

N9075A 802.16 OFDMA (WiMAX/WiBro) Measurement Application

N9076A 1xEV-DO Measurement Application

N9077A-XFP Single Acquisition Combined WLAN Measurement Application

N9079A TD-SCDMA Measurement Application

N9080A LTE FDD Measurement Application

N9082A LTE TDD Measurement Application

89601X Vector Signal Analyzer Measurement Application

Back to

“Additional Documentation” on page 87

.

N6149A-2FP iDEN/WiDEN/MotoTalk Measurement Application

• User’s & Programmer’s Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88

.

90 Chapter 2

Additional Documentation

For Specific Measurement Applications

Back to “Additional Documentation” on page 87 .

N6153A-2FP DVB-T/H Measurement Application

• User’s & Programmer’s Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88 .

Back to “Additional Documentation” on page 87 .

N6155A-2FP ISDB-T Measurement Application

• User’s & Programmer’s Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88 .

Back to “Additional Documentation” on page 87 .

N6156A-2FP DTMB Measurement Application

• User’s & Programmer’s Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88 .

Back to “Additional Documentation” on page 87 .

N6158A-2FP CMMB Measurement Application

• User’s & Programmer’s Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88 .

Back to “Additional Documentation” on page 87 .

N9020A/N9010A IQ Analyzer Measurement Application

• User’s & Programmer’s Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88 .

Back to “Additional Documentation” on page 87 .

N9020A/N9010A Spectrum Analyzer Measurement Application

• User’s & Programmer’s Reference [PDF]

Chapter 2 91

Additional Documentation

For Specific Measurement Applications

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88

.

Back to

“Additional Documentation” on page 87

.

N9061A Remote Language Compatibility Application

• Remote Language Compatibility Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88

.

Back to

“Additional Documentation” on page 87

.

N9063A Analog Demod Measurement Application

• User's and Programmer's Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88

.

Back to

“Additional Documentation” on page 87

.

N9068A Phase Noise Measurement Application

• User's and Programmer's Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88

.

Back to

“Additional Documentation” on page 87

.

N9069A Noise Figure Measurement Application

• User's and Programmer's Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88

.

Back to

“Additional Documentation” on page 87

.

N9071A GSM-EDGE Measurement Application

• User's and Programmer's Reference [PDF]

• Single Acquisition Combined GSM Measurement Application [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88

.

Back to

“Additional Documentation” on page 87

.

92 Chapter 2

Additional Documentation

For Specific Measurement Applications

N9072A cdma2000 Measurement Application

• User's and Programmer's Reference [PDF]

• Measurement Guide and Programming Examples [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88 .

Back to “Additional Documentation” on page 87 .

N9073A W-CDMA Measurement Application

• User's and Programmer's Reference [PDF]

• Single Acquisition Combined W-CDMA Measurement Application [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88 .

Back to “Additional Documentation” on page 87 .

N9074A-XFP Single Acquisition Combined Fixed WiMAX Measurement

Application

• User's and Programmer's Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88 .

Back to “Additional Documentation” on page 87 .

N9075A 802.16 OFDMA (WiMAX/WiBro) Measurement Application

• User's and Programmer's Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88 .

Back to “Additional Documentation” on page 87 .

N9076A 1xEV-DO Measurement Application

• User's and Programmer's Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88 .

Back to “Additional Documentation” on page 87 .

N9077A-XFP Single Acquisition Combined WLAN Measurement Application

• User's and Programmer's Reference [PDF]

Chapter 2 93

Additional Documentation

For Specific Measurement Applications

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88

.

Back to

“Additional Documentation” on page 87

.

N9079A TD-SCDMA Measurement Application

• User's and Programmer's Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88

.

Back to

“Additional Documentation” on page 87

.

N9080A LTE FDD Measurement Application

• User's and Programmer's Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88

.

Back to

“Additional Documentation” on page 87

.

N9082A LTE TDD Measurement Application

• User's and Programmer's Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88

.

Back to

“Additional Documentation” on page 87

.

89601X Vector Signal Analyzer Measurement Application

• User's and Programmer's Reference [PDF]

• Measurement Guide [PDF]

For details of the types of documentation available, see

“Documentation Categories” on page 88

.

Back to

“Additional Documentation” on page 87

.

94 Chapter 2

Additional Documentation

Application Notes & Other Documentation

Application Notes & Other Documentation

Click on any of the links below to open the document in a new window.

• AN-57-1: Fundamentals of RF and Microwave Noise Figure Measurements [PDF]

• AN-57-2: Noise Figure Measurement Accuracy - The Y-Factor [PDF]

• AN-150: Spectrum Analysis Basics [PDF]

• AN-1325: Performing cdma2000 Measurements Today [PDF]

• AN-1355: Designing & Testing 3GPP W-CDMA Base Transceiver Stations [PDF]

• AN-1356: Designing & Testing 3GPP W-CDMA User Equipment [PDF]

• AN-1357: Designing and Testing cdma2000 Base Stations [PDF]

• AN-1358: Designing and Testing cdma2000 Mobile Stations [PDF]

• AN-1398: Forward Link Measurements for 1xEV-DO Access Networks [PDF]

• AN-1414: Understanding Measurements of 1xEV-DO Access Terminals [PDF]

• AN-1583: Maximizing Measurement Speed with Agilent's X-Series Signal Analyzers [PDF]

• AN-1585: Using the Agilent MXA Signal Analyzer for Measuring and Troubleshooting Digitally

Modulated Signals [PDF]

• Agilent 3GPP Long Term Evolution Application Note [PDF]

• Agilent I/O Libraries - E2094Q [PDF]

• Agilent Preamplifiers and System Noise Figure Application Note [PDF]

• HSDPA and HSUPA Concepts [PDF]

• Tips for Preventing Spectrum Analyzer Damage [PDF]

• 802.16e WiMax OFDMA Signal Measurements & Troubleshooting [PDF]

• 89601X VXA Application Note Library [CHM]

• VISA User’s Guide for IO Libraries Suite

Back to Additional Documentation .

Chapter 2 95

Additional Documentation

Application Notes & Other Documentation

96 Chapter 2

3

About the Analyzer

The X-Series signal analyzer measures and monitors complex RF and microwave signals. Analog baseband analysis is available on MXA. The analyzer integrates traditional spectrum measurements with advanced vector signal analysis to optimize speed, accuracy, and dynamic range. The analyzer has

Windows XP Pro

®

built in as an operating system, which expands the usability of the analyzer.

With a broad set of applications and demodulation capabilities, an intuitive user interface, outstanding connectivity and powerful one-button measurements, the analyzer is ideal for both R&D and manufacturing engineers working on cellular, emerging wireless communications, general purpose, aerospace and defense applications.

97

About the Analyzer

Installing Application Software

Installing Application Software

When you want to install a measurement application after your initial hardware purchase, you actually only need to license it. All of the available applications are loaded in your analyzer at the time of purchase.

So when you purchase an application, you will receive an entitlement certificate that is used to obtain a license key for that particular measurement application. Enter the license key that you obtain into the

Signal Analyzer to activate the new measurement application. See below for more information.

For the latest information on Agilent Signal Analyzer measurement applications and upgrade kits, visit the following internet URL.

http://www.agilent.com/find/sa_upgrades

Viewing a License Key

Measurement personalities purchased with your instrument have been installed and activated at the factory before shipment. The instrument requires a unique License Key for every measurement application purchased. The license key is a hexadecimal string that is specific to your measurement application, instrument model number and serial number. It enables you to install, or reactivate that particular application.

Press

System

,

Show

,

System

to display which measurement applications are currently licensed in your analyzer.

Go to the following location to view the license keys for the installed measurement applications:

C:\Programing Files\Agilent\Licensing

NOTE

You may want to keep a copy of your license key in a secure location. You can print out a copy of the display showing the license numbers to do this. If you should lose your license key, call your nearest Agilent Technologies service or sales office for assistance.

Obtaining and Installing a License Key

If you purchase an additional application that requires installation, you will receive an “Entitlement

Certificate” which may be redeemed for a license key for one instrument. Follow the instructions that accompany the certificate to obtain your license key.

Installing a license key for the selected application can be done automatically using a USB memory device. To do this, you would put the license file on the USB memory device at the root level. Follow the instructions that come with your software installation kit.

Installing a license key can also be done manually using the license management application in the instrument. It is found through the instrument front panel keys at

System

,

Licensing. . .

, or internally at

C:\Programming Files\Agilent\Licensing.

98 Chapter 3

About the Analyzer

Installing Application Software

NOTE

You can also use these procedures to reinstall a license key that has been accidentally deleted, or lost due to a memory failure.

Missing and Old Measurement Application Software

All the software applications were loaded at the time of original instrument manufacture. It is a good idea to regularly update your software with the latest available version. This assures that you get any improvements and expanded functionality that is available.

Because the software was loaded at the initial purchase, there may be additional measurement applications that are now available. If the application you are interested in licensing is not available, you will need to do a software update. (Press

System

,

Show

,

System

.)

Check the Agilent internet website for the latest software versions available for downloading: http://www.agilent.com/find/pxa_software http://www.agilent.com/find/mxa_software http://www.agilent.com/find/exa_software http://www.agilent.com/find/cxa_software

You must load the updated software package into the analyzer from a USB drive, or directly from the internet. An automatic loading program is included with the files.

Chapter 3 99

About the Analyzer

X-Series Options and Accessories

X-Series Options and Accessories

“MXA Instrument Options” on page 100

“MXA Accessories” on page 101

“EXA Instrument Options” on page 101

“EXA Accessories” on page 102

“CXA Instrument Options” on page 102

“CXA Accessories” on page 103

“Advanced Measurement Application Software” on page 103

MXA Instrument Options

Product

N9020A

N9020A-503

N9020A-508

N9020A-513

N9020A-526

N9020A-B25

N9020A-BBA

N9020AK-SU1

N9020AK-PC2

N9020AK-ALL

N9020A-SSD

N9020A-HDD

N9020A-PFR

N9020A-PRC

N9020A-EA3

N9020A-EMC

N9020A-S40

N9020A-P03

N9020A-P08

N9020A-P13

Description

MXA Signal Analyzer

Frequency range from 20 Hz to 3.6 GHz

Frequency range from 20 Hz to 8.4 GHz

Frequency range from 20 Hz to 13.6 GHz

Frequency range from 20 Hz to 26.5 GHz

Analysis bandwidth, 25 MHz

Analog baseband IQ inputs

Baseband bandwidth from 25 to 40 MHz (upgrade)

Dual-Core CPU Processor with Removable Hard Drive (upgrade)

Amplitude Corrections and Limit Lines Features (upgrade)

Removable Solid State Drive Substitution

Additional Removable Hard Drive (160 GB)

Precision frequency reference

Portable configuration

Electric attenuator, 3.6 GHz

Precompliance EMI Measurement

Baseband analysis bandwidth, 40 MHz/channel

Preamplifier, 3.6 GHz

Preamplifier, 8.4 GHz

Preamplifier, 13.6 GHz

100 Chapter 3

Product

N9020A-P26

MXA Accessories

Product

N9020A-MLP

N9020A-EFM

N9020A-DVR

N9020A-MSE

N9020A-KYB

N9020A-HTC

N9020A-1CP

N9020A-1CN

N9020A-1CM

N9020A-1CR

N9020A-UK6

N9020A-1A7

N9020A-A6J

EXA Instrument Options

Product

N9010A

N9010A-503

N9010A-507

N9010A-513

N9010A-526

N9010A-B25

N9010A-FSA

N9010A-ALL

N9010A-SSD

N9010A-PC2

N9010A-HDD

Chapter 3

Description

Preamplifier, 26.5 GHz

About the Analyzer

X-Series Options and Accessories

Description

75 ohm minimum loss pad

USB flash drive

DVD-ROM drive

Mouse, USB interface

Keyboard, USB interface

Hard transit case

Rackmount kit with handles

Front handle kit

Rackmount kit

Rack slide kit

Commercial Calibration Certificate with Test Data

ISO 17025 Compliant Calibration

ANSI Z540 Compliant Calibration

Description

EXA Signal Analyzer (3.6, 7.0, 13.6, and 26.5 GHz)

Frequency range from 9 kHz to 3.6 GHz

Frequency range from 9 kHz to 7.0 GHz

Frequency range from 9 kHz to 13.6 GHz

Frequency range from 9 kHz to 26.5 GHz

Analysis bandwidth, 25 MHz

Fine step attenuator

Amplitude Corrections and Limit Lines Features (upgrade)

Removable Solid State Drive Substitution

Dual-Core Processor with Removable Hard Drive

Additional Removable Hard Drive (160 GB)

101

About the Analyzer

X-Series Options and Accessories

Product

N9010A-CPU

N9010A-PFR

N9010A-EA3

N9010A-EMC

N9010A-P03

EXA Accessories

Product

N9010A-PRC

N9010A-MLP

N9010A-EFM

N9010A-DVR

N9010A-MSE

N9010A-KYB

N9010A-HTC

N9010A-1CP

N9010A-1CN

N9010A-1CM

N9010A-1CR

N9010A-UK6

N9010A-1A7

N9010A-A6J

CXA Instrument Options

Product

N9000A

N9000A-503

N9000A-507

N9000A-PRC

N9000A-600

N9010A-FSA

102

Description

Instrument Security, Additional CPU & HDD

Precision frequency reference

Electric attenuator, 3.6 GHz

Precompliance EMI Measurement

Preamplifier, 3.6 GHz

Description

Portable Configuration

Minimum loss pad, 50 to 75 ohms

USB flash drive

DVD-ROM/CD-R/RW drive

Mouse, USB interface

Keyboard, USB interface

Hard transit case

Rackmount kit with handles

Front handle kit

Rackmount kit

Rack slide kit

Commercial Calibration Certificate with Test Data

ISO 17025 Compliant Calibration

ANSI Z540 Compliant Calibration

Description

CXA Signal Analyzer (3.0, 7.5 GHz)

Frequency range from 9 kHz to 3.0 GHz

Frequency range from 9 kHz to 7.5 GHz

Portable configuration

Benchtop configuration

Fine step attenuator

Chapter 3

About the Analyzer

X-Series Options and Accessories

Product

N9000A-EMC

N9000A-P03

N9000A-P07

CXA Accessories

Product

N9000A-MLP

N9000A-EFM

N9000A-DVR

N9000A-MSE

N9000A-KYB

N9000A-KB2

N9000A-HTC

N9000A-1CP

N9000A-1CN

N9000A-1CM

N9000A-1CR

N9000A-UK6

N9000A-1A7

N9000A-A6J

Description

EMC Basic Precompliance EMI features

Preamplifier, 3.0 GHz

Preamplifier 7.5 GHz

Description

Minimum loss pad, 50 to 75 ohms

USB flash drive

DVD-ROM/CD-R/RW drive

Mouse, USB interface

Keyboard, USB interface

Smaller keyboard, USB interface

Hard transit case

Rackmount kit with handles

Front handle kit

Rackmount kit

Rack slide kit

Commercial Calibration Certificate with Test Data

ISO 17025 Compliant Calibration

ANSI Z540 Compliant Calibration

Advanced Measurement Application Software

For a current list of application software, go to the following URLs.

For PXA,

http://www.agilent.com/find/pxa/options

Select the PXA N9030A, Options and Measurement Applications link on the top of the page.

For MXA,

http://www.agilent.com/find/mxa/options

Select the MXA N9020A, Options and Measurement Applications link on the top of the page.

For EXA,

http://www.agilent.com/find/exa/options

Chapter 3 103

About the Analyzer

X-Series Options and Accessories

Select the EXA N9010A, Options and Measurement Applications link on the top of the page.

For CXA,

http://www.agilent.com/find/cxa/options

Select the CXA N9000A, Options and Measurement Applications link on the top of the page.

104 Chapter 3

Front-Panel Features

About the Analyzer

Front-Panel Features

6

7

8

4

5

2

3

1

#

Item

Name

Menu Keys

Description

Analyzer Setup Keys

Measurement Keys

Marker Keys

Utility Keys

Probe Power

Headphones Output

Back Space Key

Key labels appear to the left of the menu keys to identify the current function of each key. The displayed functions are dependent on the currently selected

Mode and Measurement, and are directly related to the most recent key press.

These keys set the parameters used for making measurements in the current

Mode and Measurement.

These keys select the Mode, and the Measurement within the mode. They also control the initiation and rate of recurrence of measurements.

Markers are often available for a measurement, to measure a very specific point/segment of data within the range of the current measurement data

.

These keys control system-wide functionality such as:

• instrument configuration information and I/O setup,

• printer setup and printing,

• file management, save and recall,

• instrument presets.

Supplies power for external high frequency probes and accessories.

Headphones can be used to hear any available audio output.

Press this key to delete the previous character when entering alphanumeric information. It also works as the Back key in Help and Explorer windows.

Chapter 3 105

About the Analyzer

Front-Panel Features

18

19

20

9

10

#

11

12

13

14

15

16

17

Item

Description

Name

Delete Key

USB Connectors

Local/Cancel/(Esc) Key If you are in remote operation, Local:

• returns instrument control from remote back to local (the front panel).

• turns the display on (if it was turned off for remote operation).

• can be used to clear errors. (Press the key once to return to local control, and a second time to clear error message line.)

If you have not already pressed the units or Enter key, Cancel exits the currently selected function without changing its value.

Esc works the same as it does on a pc keyboard. It:

• exits Windows dialogs

• clears errors

• aborts printing

• cancels operations.

RF Input

Press this key to delete files, or to perform other deletion tasks.

Standard USB 2.0 ports, Type A. Connect to external peripherals such as a mouse, keyboard, DVD drive, or hard drive.

Numeric Keypad

Enter and Arrow Keys

Connector for inputting an external signal. Make sure that the total power of all signals at the analyzer input does not exceed +30 dBm (1 watt).

Enters a specific numeric value for the current function. Entries appear on the upper left of the display, in the measurement information area.

The Enter key terminates data entry when either no unit of measure is needed, or you want to use the default unit.

The arrow keys:

• Increment and decrement the value of the current measurement selection.

• Navigate help topics.

• Navigate, or make selections, within Windows dialogs.

• Navigate within forms used for setting up measurements.

• Navigate within tables.

The arrow keys cannot be used to move a mouse pointer around on the display.

Menu/ (Alt) Key

Ctrl Key

Select / Space Key

Tab Keys

Knob

Return Key

Alt works the same as a pc keyboard. Use it to change control focus in

Windows pull-down menus.

Ctrl works the same as a pc keyboard. Use it to navigate in Windows applications, or to select multiple items in lists.

Select is also the Space key and it has typical pc functionality. For example, in

Windows dialogs, it selects files, checks and unchecks check boxes, and picks radio button choices. It opens a highlighted Help topic.

Use these keys to move between fields in Windows dialogs.

Increments and decrements the value of the current active function.

Exits the current menu and returns to the previous menu. Has typical pc functionality.

106 Chapter 3

About the Analyzer

Front-Panel Features

Item

Description

# Name

21

22

23

24

25

26

27

28

29

30

Full Screen Key

Help Key

Speaker Control Keys

Window Control Keys

Power Standby/ On

Q Input

Q Input

I Input

I Input

Cal Out

Pressing this key turns off the softkeys to maximize the graticule display area.

Press the key again to restore the normal display.

Initiates a context-sensitive Help display for the current Mode. Once Help is accessed, pressing a front panel key brings up the help topic for that key function.

Enables you to increase or decrease the speaker volume, or mute it.

These keys select between single or multiple window displays. They zoom the current window to fill the data display, or change the currently selected window. They can be used to switch between the Help window navigation pane and the topic pane.

Turns the analyzer on. A green light indicates power on. A yellow light indicates standby mode.

The front-panel switch is a standby switch, not a LINE switch (disconnecting device). The analyzer continues to draw power even when the line switch is in standby.

The main power cord can be used as the system disconnecting device. It disconnects the mains circuits from the mains supply.

Input port for the Q channel when in differential mode.

a

Input port for the Q channel for either single or differential mode.

a

Input port for the I channel when in differential mode.

a

Input port for the I channel for either single or differential mode.

a

Output port for calibrating the I, I, Q and Q inputs and probes used with these

inputs.

a

a. Status of the LED indicates whether the current state of the port is active (green) or is not in use (dark).

Chapter 3 107

About the Analyzer

Front-Panel Features

Overview of key types

The keys labeled

FREQ Channel

,

System

, and

Marker Functions

are all examples of front-panel keys.

Most of the dark or light gray keys access menus of functions that are displayed along the right side of the display. These displayed key labels are next to a column of keys called menu keys.

Menu keys list functions based on which front-panel key was pressed last. These functions are also dependant on the current selection of measurement application (

Mode

) and measurement (

Meas

).

If the numeric value of a menu key function can be changed, it is called an active function. The function label of the active function is highlighted after that key has been selected. For example, press

AMPTD Y

Scale

. This calls up the menu of related amplitude functions. The function labeled

Ref Level

(the default selected key in the Amplitude menu) is highlighted.

Ref Level

also appears in the upper left of the display in the measurement information area. The displayed value indicates that the function is selected and its value can now be changed using any of the data entry controls.

Some menu keys have multiple choices on their label, such as

On

/

Off

or

Auto

/

Man

. The different choices are selected by pressing the key multiple times. For example, the Auto/Man type of key. To select the function, press the menu key and notice that Auto is underlined and the key becomes highlighted. To change the function to manual, press the key again so that Man is underlined. If there are more than two settings on the key, keep pressing it until the desired selection is underlined.

When a menu first appears, one key label is highlighted to show which key is the default selection. If you press

Marker Function

, the

Marker Function Off

key is the menu default key, and is highlighted.

Some of the menu keys are grouped together by a yellow bar running behind the keys near the left side or by a yellow border around the group of keys. When you press a key within the yellow region, such as

Marker Noise

, the highlight moves to that key to show it has been selected. The keys that are linked are related functions, and only one of them can be selected at any one time. For example, a marker can only have one marker function active on it. So if you select a different function it turns off the previous selection. If the current menu is two pages long, the yellow bar or border could include keys on the second page of keys.

In some key menus, a key label is highlighted to show which key has been selected from multiple available choices. And the menu is immediately exited when you press one of the other keys. For example, when you press the

Select Trace

key (in the

Trace/Detector

menu), it brings up its own menu of keys. The

Trace 1

key is highlighted. When you press the

Trace 2

key, the highlight moves to that key and the screen returns to the

Trace/Detector

menu.

If a displayed key label shows a small solid-black arrow tip pointing to the right, it indicates that additional key menus are available. If the arrow tip is not filled in solid then pressing the key the first time selects that function. Now the arrow is solid and pressing it again brings up an additional menu of settings.

108 Chapter 3

About the Analyzer

Display Annotations

Display Annotations

This section describes the display annotation as it is on the Spectrum Analyzer Measurement Application display. Other measurement application modes have some annotation differences.

Item Description

1

2

3

4

Measurement bar - Shows general measurement settings and information.

Indicates single/continuous measurement.

Some measurements include limits that the data is tested against. A Pass/Fail indication may be shown in the lower left of the measurement bar.

Active Function (measurement bar) - when the current active function has a settable numeric value, it is shown here.

Banner - shows the name of the selected application that is currently running.

Measurement title - shows title information for the current measurement, or a title that you created for the measurement.

All the keys in the Analyzer Setup part of the front panel.

Currently selected front panel key.

Mode

Meas

View/Display

,

Display

,

Title

Chapter 3 109

About the Analyzer

Display Annotations

6

7

8

Item Description

5

9

10

11

Settings panel - displays system information that is not specific to any one application.

• Input/Output status - green LXI indicates the LAN is connected. RLTS indicate Remote, Listen, Talk, SRQ

• Input impedance and coupling

• Selection of external frequency reference

• Setting of automatic internal alignment routine

Active marker frequency, amplitude or function value

Settings panel - time and date display.

Trace and detector information

Key labels that change based on the most recent key press.

Displays information, warning and error messages. Message area - single events, Status area - conditions

Measurement settings for the data currently being displayed in the graticule area. In the example above: center frequency, resolution bandwidth, video bandwidth, frequency span, sweep time and number of sweep points.

Local

and

System, I/O Config

Input/Output

,

Amplitude

,

System

and others

Marker

System

,

Control Panel

Trace/Detector

,

Clear Write (W) Trace

Average (A) Max Hold (M) Min Hold

(m)

Trace/Detector

,

More, Detector,

Average (A) Normal (N) Peak (P)

Sample (S) Negative Peak (p)

Softkeys

Keys in the Analyzer Setup part of the front panel.

110 Chapter 3

Rear-Panel Features

MXA and EXA with Option PC2

EXA

CXA

Chapter 3

About the Analyzer

Rear-Panel Features

111

About the Analyzer

Rear-Panel Features

1

#

Item

Name

EXT REF IN

Description

Input for an external frequency reference signal:

2

3

4

GPIB

For MXA – 1 to 50 MHz

For EXA – 10 MHz.

For CXA – 10 MHz.

A General Purpose Interface Bus (GPIB, IEEE 488.1) connection that can be used for remote analyzer operation.

USB Connector USB 2.0 port, Type B. USB TMC (test and measurement class) connects to an external pc controller to control the instrument and for data transfers over a 480 Mbps link.

USB Connectors Standard USB 2.0 ports, Type A. Connect to external peripherals such as a mouse, keyboard, printer, DVD drive, or hard drive.

7

8

5

6

MONITOR

LAN

Allows connection of an external VGA monitor.

A TCP/IP Interface that is used for remote analyzer operation.

Line power input The AC power connection. See the product specifications for more details.

Removable Disk

Drive

Standard on MXA. Optional on EXA.

9 Digital Bus

10 Analog Out

Reserved for future use.

For PXA option YAV:

Screen Video

Log Video

Linear Video

Demod Audio

11 TRIGGER 2

OUT

12 TRIGGER 1

OUT

13 Sync

14 TRIGGER 2 IN

15 TRIGGER 1 IN

16 Noise Source

Drive +28 V

(Pulsed)

17 SNS Series Noise

Source

A trigger output used to synchronize other test equipment with the analyzer. Configurable from the Input/Output keys.

A trigger output used to synchronize other test equipment with the analyzer. Configurable from the Input/Output keys.

Reserved for future use.

Allows external triggering of measurements.

Allows external triggering of measurements.

For use with Agilent 346A, 346B, and 346C Noise Sources.

For use with Agilent N4000A, N4001A, N4002A Smart Noise Sources

(SNS).

112 Chapter 3

About the Analyzer

Rear-Panel Features

Item

Description

# Name

18 10 MHz OUT

19 Preselector Tune

Out

20 Aux IF Out

An output of the analyzer internal 10 MHz frequency reference signal. It is used to lock the frequency reference of other test equipment to the analyzer.

Reserved for future use.

For PXA options:

CR3 Second IF Out

CRP Arbitrary IF Out

ALV Log Video

Chapter 3 113

About the Analyzer

Window Control Keys

Window Control Keys

The instrument provides three front-panel keys for controlling windows. They are

Multi Window

,

Zoom

, and

Next Window

. These are all “immediate action” keys.

Multi-Window

The

Multi Window

front-panel key is not used at this time. It is there to support future functionality.

Key Path

Initial S/W Revision

Front-panel key

Prior to A.02.00

Zoom

Zoom is a toggle function. Pressing once Zooms the selected window; pressing again un-zooms.

When Zoom is on for a window, that window will get the entire primary display area. The zoomed window, since it is the selected window, is outlined in green.

Zoom is local to each Measurement. Each Measurement remembers its Zoom state. The Zoom state of each Measurement is part of the Mode’s state.

NOTE

Remote Command:

Remote Command:

Example:

Preset:

Data acquisition and processing for the other windows continues while a window is zoomed, as does all SCPI communication with the other windows.

:DISPlay:WINDow:FORMat:ZOOM

:DISPlay:WINDow:FORMat:TILE

:DISP:WIND:FORM:ZOOM sets zoomed

:DISP:WIND:FORM:TILE sets un-zoomed

TILE

114 Chapter 3

About the Analyzer

Window Control Keys

Initial S/W Revision: Prior to A.02.00

Next Window

Selects the next window of the current view.

When this key is selected in Help Mode, it toggles focus between the table of contents window and the topic pane window.

Remote Command:

Example:

Preset:

Min:

Max:

Initial S/W Revision:

:DISPlay:WINDow[:SELect] <number>

:DISPlay:WINDow[:SELect]?

:DISP:WIND 1

1

1

If <number> is greater than the number of windows, limit to <number of windows>

Prior to A.02.00

Selected Window

One and only one window is always selected. The selected window has the focus and all key presses are going to that window.

The selected window has a green boundary. If a window is not selected, its boundary is gray.

If a window in a multi-window display is zoomed it is still outlined in green. If there is only one window, the green outline is not used. This allows the user to distinguish between a zoomed window and a display with only one window.

The selected window is local to each Measurement. Each Measurement remembers which window is selected. The selected window for each Measurement is remembered in Mode state.

Navigating Windows

When the Next Window key is pressed, the next window in the order of precedencebecomes selected. If the selected window was zoomed, the next window will also be zoomed.

The window navigation does NOT use the arrow and select keys. Those are reserved for navigation within a window.

Chapter 3 115

About the Analyzer

Mouse and Keyboard Control

Mouse and Keyboard Control

If you do not have access to the instrument front-panel, there are several ways that a mouse and PC

Keyboard can give you access to functions normally accessed using the front-panel keys.

Right-Click

If you plug in a mouse and right-click on the analyzer screen, a menu will appear as below:

Placing the mouse on one of the rows marked with a right arrow symbol will cause that row to expand, as for example below where the mouse is hovered over the “Utility” row:

116 Chapter 3

About the Analyzer

Mouse and Keyboard Control

This method can be used to access any of the front-panel keys by using a mouse; as for example if you are accessing the instrument through Remote Desktop.

The array of keys thus available is shown below:

Utility

Window

Return

Local

Help

Exit

Setup

Marker

Control

Mode

Meas

Mode Preset

User Preset

(list of

Measurements)

Zoom

Split Screen

Next

Full Screen

(list of Modes)

Frequency

File

Quick Save

Save

Recall

Page Setup

Print

BW

Trigger

Sweep-Control

Restart

Single

Continuous

Marker

Peak Search

Marker To

Marker Function

Frequency

Span

Amplitude

Input-Output

View-Display

Trace-Detector

Auto Couple

BW

Source

Mode Setup

Meas Setup

PC Keyboard

If you have a PC keyboard plugged in (or via Remote Desktop), certain key codes on the PC keyboard

Chapter 3 117

About the Analyzer

Mouse and Keyboard Control

map to front-panel keys on the GPSA front panel. These key codes are shown below:

Save

Recall

Mode Preset

User Preset

Print

File

Mode

Measure

Mode Setup

Meas Setup

Trigger

Sweep/Control

Restart

Single

Front-panel key

Frequency

Span

Amplitude

Input/Output

View/Display

Trace/Detector

Auto Couple

Bandwidth

Source

Marker

Peak Search

Marker To

Marker Function

System

Quick Save

CTRL+S

CTRL+R

CTRL+M

CTRL+U

CTRL+P

CTRL+SHIFT+L

CTRL+SHIFT+M

CTRL+ALT+M

CTRL+SHIFT+E

CTRL+ALT+E

CTRL+SHIFT+T

CTRL+SHIFT+W

CTRL+ALT+R

CTRL+ALT+S

Key code

CTRL+SHIFT+F

CTRL+SHIFT+S

CTRL+SHIFT+A

CTRL+SHIFT+O

CTRL+SHIFT+V

CTRL+ALT+T

CTRL+SHIFT+C

CTRL+ALT+B

CTRL+SHIFT+E

CTRL+ALT+K

CTRL+ALT+P

CTRL+ALT+N

CTRL+ALT+F

CTRL+SHIFT+Y

CTRL+Q

118 Chapter 3

Select

Up Arrow

Down Arrow

Left Arrow

Right Arrow

Menu key 1

Menu key 2

Menu key 3

Menu key 4

Menu key 5

Menu key 6

Menu key 7

Backspace

Enter

Tab

1

Dec Audio

Help

Control

Alt

Enter

Cancel

Del

Backspace

Cont

Zoom

Next Window

Split Screen

Full Screen

Return

Mute

Inc Audio

Chapter 3

Space

Up

Down

Left

Right

CTRL+SHIFT+F1

CTRL+SHIFT+F2

CTRL+SHIFT+F3

CTRL+SHIFT+F4

CTRL+SHIFT+F5

CTRL+SHIFT+F6

CTRL+SHIFT+F7

BACKSPACE

ENTER

Tab

1

CTRL+ALT+C

CTRL+SHIFT+Z

CTRL+SHIFT+N

CTRL+L

CTRL+SHIFT+B

CTRL+SHIFT+R

Mute

Volume Up

Volume Down

F1

CTRL

ALT

Return

Esc

Delete

Backspace

About the Analyzer

Mouse and Keyboard Control

119

About the Analyzer

Mouse and Keyboard Control

8

9

6

7

0

4

5

2

3

Here is a pictorial view of the table:

8

9

6

7

0

4

5

2

3

120 Chapter 3

About the Analyzer

Instrument Security & Memory Volatility

Instrument Security & Memory Volatility

If you are using the instrument in a secure environment, you may need details of how to clear or sanitize its memory, in compliance with published security standards of the United States Department of

Defense, or other similar authorities.

For the X Series analyzers, this information is contained in the document "Security Features and

Volatility". This document is not included in the Documentation CD, or the instrument’s on-disk library, but it may be downloaded from Agilent’s web site.

To obtain a copy of the document, click on or browse to the following URL: http://www.agilent.com/find/security

To locate and download the document, select Model Number "N9020A", then click "Submit". Then, follow the on-screen instructions to download the file.

Chapter 3 121

About the Analyzer

Instrument Security & Memory Volatility

122 Chapter 3

4

About the Agilent 89601X VXA Signal

Analyzer Measurement

This chapter provides overall information on the Agilent 89601X VXA Signal Analyzer Measurement

Application and describes the measurements made by the analyzer. Installation instructions for adding this option to your analyzer are provided in this section, in case you purchased this option separately.

123

About the Agilent 89601X VXA Signal Analyzer Measurement

What Does the Agilent 89601X VXA Signal Analyzer Measurement Application Do?

What Does the Agilent 89601X VXA Signal Analyzer

Measurement Application Do?

The 89601X VXA is a full-featured vector signal analyzer that can help determine if an RF modulated source or transmitter is working correctly. There are standard and optional measurements for complete analysis and demodulation of most communications signals.

The Vector Analysis measurement (included in Option 205) provides:

• Spectrum analysis and Time Domain analysis with Signal Tracking

• Band Power, Occupied Bandwidth, and ACP measurements

• Markers, Marker Coupling, Triggering

• Time Gating

• Frequency Counter

The Analog Demodulation measurement (included in Option 205) will perform all the above and will also analyze and demodulate signals that use the following modulation formats:

• AM, FM, PM

The Digital Demodulation measurement (Option AYA) will analyze and demodulate signals that use the following modulation formats:

• MSK

• QPSK

• 8PSK

• BPSK

π/4 DQPSK

• DQPSK

π/8 D8PSK

• D8PSK

• Offset QPSK

• QAM16, 32, 64, 128, 256, 512, 1024

• DVB QAM 16, 32, 64, 128, 256

• FSK 2, 4, 8, 16 states

• VSB8, VSB16

• APSK16, 32, 16 w/dub, 32 w/dub

124 Chapter 4

About the Agilent 89601X VXA Signal Analyzer Measurement

What Does the Agilent 89601X VXA Signal Analyzer Measurement Application Do?

• Standard communications formats provided by Option AYA include:

— Cellular: IS–95 Base and Mobile, GSM, EDGE, CDPD, NADC, PDC, PHP, 3GPP (W-CDMA)

— Wireless Networking: 802.11b, HIPERLAN/1 (HBR and LBR), Bluetooth, ZigBee 868, 915, and

2450

— Digital Video: DTV8, DTV16, DVB16, DVB32, DVB64, DVB128, DVB256, DVB 16APSK with code rates 2/3 to 9/10, DVB 32 APSK with code rates 3/4 to 9/10.

— Other: APCO 25, DECT, TETRA, VDL Mode 3

With other digital demod options, the 89601X VXA will also demodulate more complicated signals that conform to the following standard communications formats:

• Option B7R - WLAN 802.11 a/b/g, OFDM and DSSS

• Option H09 - iDEN, WiDEN, and MotoTalk

NOTE

For CDMA signals, 89601X VXA can analyze digital modulation for a single code channel only. If multiple code channels are transmitted, synchronization will fail, and incorrect EVM results will be obtained. For modulation quality measurements of multiple code channels, Modulation Accuracy and Code Domain measurements must be performed by a full-featured standard-based measurement application, like N9073for W-CDMA.

Chapter 4 125

About the Agilent 89601X VXA Signal Analyzer Measurement

Installing Application Software

Installing Application Software

When you want to install a measurement application after your initial hardware purchase, you actually only need to license it. All of the available applications are loaded in your analyzer at the time of purchase.

So when you purchase an application, you will receive an entitlement certificate that is used to obtain a license key for that particular measurement application. Enter the license key that you obtain into the

N9020A Signal Analyzer to activate the new measurement application. See below for more information.

For the latest information on Agilent Signal Analyzer measurement applications and upgrade kits, visit the following internet URL.

http://www.agilent.com/find/sa_upgrades

Viewing a License Key

Measurement personalities purchased with your instrument have been installed and activated at the factory before shipment. The instrument requires a unique License Key for every measurement application purchased. The license key is a hexadecimal string that is specific to your measurement application, instrument model number and serial number. It enables you to install, or reactivate that particular application.

Press

System

,

Show

,

System

to display which measurement applications are currently licensed in your analyzer.

Go to the following location to view the license keys for the installed measurement applications:

C:\Programing Files\Agilent\Licensing

NOTE

You may want to keep a copy of your license key in a secure location. You can print out a copy of the display showing the license numbers to do this. If you should lose your license key, call your nearest Agilent Technologies service or sales office for assistance.

Obtaining and Installing a License Key

If you purchase an additional application that requires installation, you will receive an “Entitlement

Certificate” which may be redeemed for a license key for one instrument. Follow the instructions that accompany the certificate to obtain your license key.

Installing a license key for the selected application can be done automatically using a USB memory device. To do this, you would put the license file on the USB memory device at the root level. Follow the instructions that come with your software installation kit.

Installing a license key can also be done manually using the license management application in the instrument. It is found through the instrument front panel keys at

System

,

Licensing. . .

, or internally at

C:\Programming Files\Agilent\Licensing.

126 Chapter 4

About the Agilent 89601X VXA Signal Analyzer Measurement

Installing Application Software

NOTE

You can also use these procedures to reinstall a license key that has been accidentally deleted, or lost due to a memory failure.

Missing and Old Measurement Application Software

All the software applications were loaded at the time of original instrument manufacture. It is a good idea to regularly update your software with the latest available version. This assures that you get any improvements and expanded functionality that is available.

Because the software was loaded at the initial purchase, there may be additional measurement applications that are now available. If the application you are interested in licensing is not available, you will need to do a software update. (Press

System

,

Show

,

System

.)

Check the Agilent internet website for the latest software versions available for downloading: http://www.agilent.com/find/mxa_software http://www.agilent.com/find/exa_software

You must load the updated software package into the analyzer from a USB drive, or directly from the internet. An automatic loading program is included with the files.

Chapter 4 127

About the Agilent 89601X VXA Signal Analyzer Measurement

Installing Application Software

128 Chapter 4

5

System Functions

129

System Functions

File

File

Opens a menu that enables you to access various standard and custom Windows functions. Press any other front-panel key to exit

Key Path

Initial S/W Revision

Front-panel key

Prior to A.02.00

File Explorer

Opens the standard Windows File Explorer. The File Explorer opensin the My Documents directory for the current user.

The File Explorer is a separate Windows application, so to return to the analyzer once you are in the File

Explorer, you may either:

Exit the File Explorer by clicking on the red X in the upper right hand corner, with a mouse

Or use Alt-Tab: press and hold the Alt key and press and release the Tab key until the Analyzer

130

System Functions

File

logo is showing in the window in the center of the screen, as above, then release the Alt key.

Key Path

Initial S/W Revision

File

Prior to A.02.00

Page Setup

The Page Setup key brings up a Windows Page Setup dialog that allows you to control aspects of the pages sent to the printer when the PRINT hardkey is pressed.

Key Path

Initial S/W Revision

File

Prior to A.02.00

Paper size, the printer paper source, the page orientation and the margins are all settable. Just like any standard Windows dialog, you may navigate the dialog using front-panel keys, or a mouse. There are no

SCPI commands for controlling these parameters.

Also contained in this dialog is a drop-down control that lets you select the Theme to use when printing.

For more on Themes, see information under View/Display, Display, System Display Settings, Theme.

The Theme control has a corresponding SCPI command:

Parameter Name

Parameter Type

Mode

Remote Command

Example

Print Themes

Enum

All

:SYSTem:PRINt:THEMe

TDColor|TDMonochrome|FCOLor|FMONochrome

:SYSTem:PRINt:THEMe?

:SYST:PRIN:THEM FCOL

131

System Functions

File

Setup

Preset

State Saved

Initial S/W Revision

:SYSTem:DEFault MISC

FCOL; not part of Preset, but is reset by Restore Misc Defaults or Restore

System Defaults All and survives subsequent running of the modes.

No

Prior to A.02.00

Print

Refer to your Microsoft Windows Operating System manual.

Exit

This key, when pressed, will exit the Instrument Application. A dialog box is used to confirm that you intended to exit the application:

Key Path

Mode

Notes

Initial S/W Revision

File

All

The Instrument Application will close. No further SCPI commands can be sent. Use with caution!

Prior to A.02.00

132

System Functions

Preset

Preset

Mode Preset

Returns the active mode to a known state.

Mode Preset

does the following for the currently active mode:

• Aborts the currently running measurement.

• Brings up the default menu for the mode, with no active function.

• Sets measurement settings to their preset values for the active mode only.

• Activates the default measurement.

• Brings up the default menu for the mode.

• Clears the input and output buffers.

• Sets Status Byte to 0.

Mode Preset does not:

Cause a mode switch

Affect mode persistent settings

Affect system settings

Key Path

Remote Command

Example

Notes

Couplings

Front-panel key

:SYSTem:PRESet

:SYST:PRES

*RST is preferred over :SYST:PRES for remote operation. *RST does a Mode

Preset, as done by the :SYST:PRES command, and sets the measurement mode to Single measurement rather than Continuous for optimal remote control throughput.

Clears all pending OPC bits. The Status Byte is set to 0.

A Mode Preset aborts the currently running measurement, activates the default measurement, and. gets the mode to a consistent state with all of the default couplings set.

Prior to A.02.00

Initial S/W Revision

How-To Preset

The table below shows all possible presets, their corresponding SCPI commands and front-panel access

(key paths). Instrument settings depend on the current measurement context. Some settings are local to the current measurement, some are global (common) across all the measurements in the current mode, and some are global to all the available modes. In a similar way, restoring the settings to their preset state can be done within the different contexts.

133

System Functions

Preset

Auto Couple - is a measurement local key. It sets all Auto/Man parameter couplings in the

measurement to Auto. Any Auto/Man selection that is local to other measurements in the mode will not be affected.

Meas Preset - is a measurement local key. Meas Preset resets all the variables local to the current

measurement except the persistent ones.

Mode Preset - resets all the current mode's measurement local and measurement global variables

except the persistent ones.

Restore Mode Defaults - resets ALL the Mode variables (and all the Meas global and Meas local

variables), including the persistent ones.

Type Of Preset

Auto Couple

Meas Preset

Mode Preset

Restore Mode Defaults

Restore All Mode Defaults

*RST

Restore Input/Output

Defaults

Restore Power On Defaults

Restore Alignment Defaults

SCPI Command

:COUPle ALL

:CONFigure:<Measurement>

:SYSTem:PRESet

:INSTrument:DEFault

:SYSTem:DEFault MODes

*RST

:SYSTem:DEFault INPut

:SYSTem:DEFault PON

:SYSTem:DEFault ALIGn

Restore Miscellaneous

Defaults

:SYSTem:DEFault MISC

Restore All System Defaults :SYSTem:DEFault [ALL]

:SYSTem:PRESet:PERSistent

User Preset

User Preset All Modes

Power On Mode Preset

Power On User Preset

Power On Last State

:SYSTem:PRESet:USER

:SYSTem:PRESet:USER:ALL

:SYSTem:PON:TYPE MODE

:SYSTem:PON:TYPE USER

:SYSTem:PON:TYPE LAST

Front Panel Access

Auto Couple front-panel key

Meas Setup Menu

Mode Preset (green key)

Mode Setup Menu

System Menu; Restore

System Default Menu not possible (Mode Preset with Single)

System Menu; Restore

System Default Menu

System Menu; Restore

System Default Menu

System Menu; Restore

System Default Menu

System Menu; Restore

System Default Menu

System Menu; Restore

System Default Menu

User Preset Menu

User Preset Menu

System Menu

System Menu

System Menu

Restore Mode Defaults

Resets the state for the currently active mode by resetting the mode persistent settings to their factory

134

System Functions

Preset

default values, clearing mode data and by performing a Mode Preset. This function will never cause a mode switch. This function performs a full preset for the currently active mode; whereas, Mode Preset performs a partial preset. Restore Mode Defaults does not affect any system settings. System settings are reset by the Restore System Defaults function. This function does reset mode data; as well as settings.

Key Path

Remote Command

Example

Notes

Couplings

Initial S/W Revision

Mode Setup

:INSTrument:DEFault

:INST:DEF

Clears all pending OPC bits. The Status Byte is set to 0.

A message comes up saying: “If you are sure, press key again”.

A Restore Mode Defaults will cause the currently running measurement to be aborted and causes the default measurement to be active. It gets the mode to a consistent state with all of the default couplings set.

Prior to A.02.00

*RST (Remote Command Only)

*RST is equivalent to :SYST:PRES;:INIT:CONT OFF which is a Mode Preset in the Single measurement state. This remote command is preferred over Mode Preset remote command -

:SYST:PRES, as optimal remote programming occurs with the instrument in the single measurement state.

Remote Command:

Example:

Notes:

Couplings:

Initial S/W Revision:

*RST

*RST

Sequential

Clears all pending OPC bits and the Status Byte is set to 0.

A *RST will cause the currently running measurement to be aborted and cause the default measurement to be active. *RST gets the mode to a consistent state with all of the default couplings set.

Prior to A.02.00

135

System Functions

Print

Print

This front-panel key is equivalent to performing a File, Print, OK. It immediately performs the currently configured Print to the Default printer.

The :HCOPy command is equivalent to pressing the PRINT key. The HCOPy:ABORt command can be used to abort a print which is already in progress. Sending HCOPy:ABORt will cause the analyzer to stop sending data to the printer, although the printer may continue or even complete the print, depending on how much data was sent to the printer before the user sent the ABORt command.

Key Path

Remote Command

Initial S/W Revision

Front-panel key

:HCOPy[:IMMediate]

Prior to A.02.00

Key Path

Remote Command

Initial S/W Revision

SCPI command only

:HCOPy:ABORt

Prior to A.02.00

136

System Functions

Quick Save

Quick Save

The Quick Save front-panel key repeats the most recent save that was performed from the Save menu, with the following exceptions: :

Register saves are not remembered as Saves for the purpose of the Quick Save function

If the current measurement does not support the last non-register save that was performed, an informational message is generated, “File type not supported for this measurement”

Quick Save repeats the last type of qualified save (that is, a save qualified by the above criteria) in the last save directory by creating a unique filename using the Auto File Naming algorithm described below.

If Quick Save is pressed after startup and before any qualified Save has been performed, the Quick Save performs a Screen Image save using the current settings for Screen Image saves (current theme, current directory), which then becomes the “last save” for the purpose of subsequent Quick Saves.

The Auto File Naming feature automatically generates a file name for use when saving a file. The filename consists of a prefix and suffix separated by a dot, as is standard for the Windows® file system.

A default prefix exists for each of the available file types:

Type

State

Trace + State

Screen

Default Prefix

State_

State_

Screen_

Amplitude Corrections Ampcor_

Traces Trace_

Limit Lines

Measurement Result

Capture Buffer

LLine_

MeasR_

CapBuf_

Menu

(Save/Recall)

(Save/Recall)

(Save/Recall)

(Import/Export)

(Import/Export)

(Import/Export)

(Import/Export)

(Import/Export)

A four digit number is appended to the prefix to create a unique file name. The numbering sequence starts at 0000 within each Mode for each file type and works its way up to 9999, then wraps to 0000 again. It remembers where it was through a Mode Preset and when leaving and returning to the Mode. It is reset by Restore Misc Defaults and Restore System Defaults and subsequent running of the instrument application. So, for example, the first auto file name generated for State files is State_0000.state. The next is State_0001, and so forth.

One of the key features of Auto File Name is that we guarantee that the Auto File Name will never conflict with an existing file. The algorithm looks for the next available number. If it gets to 9999, then it looks for holes. If it find no holes, that is no more numbers are available, it gives an error.

For example, if when we get to State_0010.state there is already a State_0010.state file in the current directory, it advances the counter to State_0011.state to ensure that no conflict will exist (and then it verifies that State_0011.state also does not exist in the current directory and advances again if it does, and so forth).

137

.

System Functions

Quick Save

If you enter a file name for a given file type, then the prefix becomes the filename you entered instead of the default prefix, followed by an underscore. The last four letters (the suffix) are the 4-digit number.

For example, if you save a measurement results file as “fred.csv”, then the next auto file name chosen for a measurement results save will be fred_0000.csv.

NOTE

Although 0000 is used in the example above, the number that is used is actually the current number in the Meas Results sequence, that is, the number that would have been used if you had not entered your own file name.

NOTE

If the filename you entered ends with _dddd, where d=any number, making it look just like an auto file name, then the next auto file name picks up where you left off with the suffix being dddd + 1.

Key Path

Notes

Initial S/W Revision

Front-panel key

No remote command for this key specifically.

Prior to A.02.00

138

System Functions

Recall

Recall

Accesses a menu that enables you to select the information that you want to recall.

The options are State, Trace and Data. (screen images can be saved, but not recalled.) The default paths for Recall are data type dependent and are the same as for the Save key.

Key Path

Notes

Initial S/W Revision

Front-panel key

No remote command directly controls the Recall Type that this key controls.

The Recall type is a node in the :MMEM:LOAD command. An example is

:MMEM:LOAD:STATe <filename>.

Prior to A.02.00

State

Accesses a menu that enables you to recall a State that has previously been saved. Recalling a saved state returns the analyzer as close as possible to the mode context and may cause a mode switch if the file selected is not for the current active mode. A State file can be recalled from either a register or a file.

Once you select the source of the recall in the State menu, the recall will occur.

Key Path

Mode

Example

Notes

Initial S/W Revision

Recall

All

MMEM:LOAD:STAT "MyStateFile.state"

This loads the state file data (on the default file directory path) into the instrument state.

See

“Open” on page 141 .

Prior to A.02.00

In measurements that support saving Traces, for example, Swept SA, the Trace data is saved along with the State in the State file. When recalling the State, the Trace data is recalled as well. Traces are recalled exactly as they were stored, including the writing mode and update and display modes. If a Trace was updating and visible when the State was saved, it will come back updating and visible,and its data will be rewritten right away. When you use State to save and recall traces, any trace whose data must be preserved should be placed in View or Blank mode before saving.

The following table describes the Trace Save and Recall possibilities:

You want to recall state and one trace’s data, leaving other traces unaffected.

Save Trace+State from 1 trace.

Make sure that no other traces are updating (they should all be in

View or Blank mode) when the save is performed.

On Recall, specify the trace you want to load the one trace’s data into. This trace will load in View.

All other traces’ data will be unaffected, although their trace mode will be as it was when the state save was performed.

139

System Functions

Recall

You want to recall all traces Save Trace+State from ALL traces.

You want all traces to load exactly as they were when saved.

Save State

On Recall, all traces will come back in View (or Blank if they were in Blank or Background when saved)

On recall, all traces’ mode and data will be exactly as they were when saved. Any traces that were updating will have their data immediately overwritten.

Register 1 thru Register 6

Selecting any one of these register keys causes the State of the mode from the specified Register to be recalled. Each of the register keys annotates whether it is empty or at what date and time it was last modified.

Registers are shared by all modes, so recalling from any one of the 6 registers may cause a mode switch to the mode that was active when the save to the Register occurred.

After the recall completes, the message "Register <register number> recalled" appears in the message bar.

Key Path

Example

Readback

Initial S/W Revision

Recall, State

*RCL 1

Date and time with seconds resolution of the last Save is displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Key Path

Example

Readback

Initial S/W Revision

Key Path

Example

Readback

Initial S/W Revision

Key Path

Example

Recall, State

*RCL 2

Date and time with seconds resolution of the last Save is displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Recall, State

*RCL 3

Date and time with seconds resolution of the last Save is displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Recall, State

*RCL 4

140

System Functions

Recall

Readback

Initial S/W Revision

Key Path

Example

Readback

Initial S/W Revision

Date and time with seconds resolution of the last Save is displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Recall, State

*RCL 5

Date and time with seconds resolution of the last Save is displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Key Path

Example

Readback

Recall, State

*RCL 6

Date and time with seconds resolution of the last Save is displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Initial S/W Revision

From File\ File Open

Brings up the standard Windows® File Open dialog and its corresponding key menu.

When you first enter this dialog, the State File default path is in the Look In: box in this File Open dialog.

The File Open dialog is loaded with the file information related to the State Save Type. The first *.state file is highlighted. The only files that are visible are the *.state files and the Files of type is *.state, since

.state is the file suffix for the State save type. For more details, refer to “Open” on page 151

.

Key Path

Notes

Initial S/W Revision

Recall, State

Brings up the Open dialog for recalling a State Save Type

Prior to A.02.00

Open

The recalling State function must first verify the file is recallable in the current instrument by checking the software version and model number of the instrument. If everything matches, a full recall proceeds by aborting the currently running measurement, and then loading the State from the saved state file to as close as possible to the context in which the save occurred. You can open state files from any mode, so recalling a State file switches to the mode that was active when the save occurred. After switching to the mode of the saved state file, mode settings and data (if any for the mode) are loaded with values from the saved file. The saved measurement of the mode becomes the newly active measurement and the data relevant to the measurement (if there is any) is recalled.

If there is a mismatch between file version or model number or instrument version or model number, the recall function tries to recall as much as possible and it returns a warning message of what it did.

NOTE

No Trace data is loaded when recalling a State File. Measurements that support

141

System Functions

Recall

loading of trace data will include a Trace key in the Recall menu and will load

State + Trace data from .trace files under that key.

Key Path

Remote Command

Example

Notes

Notes

Recall, State, From File…

:MMEMory:LOAD:STATe <filename>

:MMEM:LOAD:STAT "myState.state" recalls the file myState.state on the default path

Auto return to the State menu and the Open dialog goes away.

Advisory Event "Recalled File <file name>" after recall is complete.

If the file specified is empty an error is generated. If the specified file does not exist, another error is generated. If there is a mismatch between the file and the proper file type, an error is generated. If there is a mismatch between file version or model number or instrument version or model number, a warning is displayed. Then it returns to the State menu and File Open dialog goes away.

Although the trace data is included in the .state file it is not recalled. Recalling trace data is left for .trace files only for measurements that support recalling of trace data. Errors are generated if the specified file is empty or does not exist, or there is a file type mismatch.

Prior to A.02.00

Initial S/W Revision

The state of a mode includes all of the variables affected by doing a full preset. It not only recalls Mode

Preset settings, but it also recalls all of the mode persistent settings and data if the mode has either. Each mode determines whether data is part of mode state and if the mode has any persistent settings.

Recall

State

also recalls all of the

Input/Output

system settings, since they are saved with each State File for each mode.

The Recall State function does the following:

Verifies that the file is recallable on this instrument using the version number and model number.

Aborts the currently running measurement.

Clears any pending operations.

Switches to the mode of the selected Save State file.

Sets mode State and Input/Output system settings to the values in the selected Saved State file.

Limits settings that differ based on model number, licensing or version number.

Makes the saved measurement for the mode the active measurement.

Clears the input and output buffers.

Status Byte is set to 0.

Executes a *CLS

142

System Functions

Recall

Trace (+State)

Select Trace as the data type to be recalled. Trace files include the state of the mode they were saved from as well as the trace data, with internal flags to indicate which trace the user was trying to save, which may include ALL traces. They are otherwise identical to State files. Recalling

trace data

may cause a mode switch if the file selected is not for the currently active mode.

Not all modes support saving of trace data with the state, and for modes that do, not all measurements do.

The Trace key is grayed out for measurements that do not support trace recall. It is blanked for modes that do not support trace recall.

This key will not actually cause the recall, since the recall feature still needs to know from which file to recall the trace and which trace to recall it into. Pressing this key will bring up the Recall Trace menu that provides you with the options of where to retrieve the trace.

For quick recalls, the Trace menu lists 5 registers to recall from or you can select a file to recall from.

Key Path

Mode

Example

Initial S/W Revision

Recall

SA

MMEM:LOAD:TRAC TRACE2,"MyTraceFile.trace"

This loads the trace file data (on the default file directory path) into the specified trace.

:MMEM:LOAD:TRAC:REG TRACE1,2 restores the trace data in register 2 to Trace 1

Prior to A.02.00

Register 1 thru Register 5

Selecting any one of these register keys causes the Traces and State from the specified Register to be recalled. Each of the register keys annotates whether it is empty or at what date and time it was last modified.

Trace registers are shared by all modes, so recalling from any one of the 5 registers may cause a mode switch to the mode that was active when the save to the Register occurred.

After the recall completes, the message “Trace Register <register number> recalled” appears in the message bar.

Key Path

Readback

Initial S/W Revision

Recall, Trace

Date and time with seconds resolution of the last Save is displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Key Path

Readback

Recall, Trace

Date and time with seconds resolution of the last Save is displayed on the key, or "(empty)" if no prior save operation performed to this register.

143

System Functions

Recall

Initial S/W Revision

Key Path

Readback

Initial S/W Revision

Key Path

Readback

Initial S/W Revision

Prior to A.02.00

Recall, Trace

Date and time with seconds resolution of the last Save is displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Recall, Trace

Date and time with seconds resolution of the last Save is displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Key Path

Readback

Recall, Trace

Date and time with seconds resolution of the last Save is displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Initial S/W Revision

To Trace

These menu selections let you pick which Trace to recall the saved trace into. Not all modes have the full

6 traces available. The default is the currently selected trace, selected in this menu or in the

Trace/Detector, Export Data, Import Data, or Save Trace menus, except if you have chosen All, then it remains chosen until you specifically change it to a single trace.

If the .trace file is an "all trace" file, "

To Trace

" is ignored and the traces each go back to the trace they were saved from.

Once selected, the key returns back to the Recall Trace menu and the selected Trace number is annotated on the key. Now you have selected exactly where the trace needs to be recalled. To trigger a recall of the selected Trace, you must select the

Open

key in the Recall Trace menu.

Key Path

Mode

Initial S/W Revision

Save, Data, Trace

SA

Prior to A.02.00

Open…

Accesses the standard Windows File Open dialog and its corresponding File Open menu. When you navigate to this selection, you have already determinedyou are recalling Trace and now you want to specify from which file to do the recall.

When you first enter this dialog, the State File default path is in the Look In: box. The

File Open

dialog is loaded with the file information related to the State Save Type. The first *.trace file is highlighted.

Also, the only files that are visible are the *.trace files and the Files of type is *.trace, since .trace is the

144

System Functions

Recall

file suffix for the Trace save type. For more details, refer to “File Open Dialog and Menu” on page 150

.

Key Path

Mode

Notes

Initial S/W Revision

Recall, Trace

SA

Brings up Open dialog for recalling a Trace Save Type

Prior to A.02.00

Open

The recalling Trace function must first verify the file is recallable in this instrument by checking instrument software version and model number, since it includes State. If everything matches, a full recall proceeds by aborting the currently running measurement, and loading the state from the saved state file to as close as possible to the context in which the save occurred. You can open .trace files from any mode that supports them, so recalling a Trace file switches to the mode that was active when the save occurred. After switching to the mode of the saved state file, mode settings and data (if any for the mode) are loaded with values from the saved file and the saved measurement of the mode becomes the newly active measurement, and the data relevant to the measurement (if there is any) is recalled.

Once the state is loaded, the trace data must be loaded. The internal flags are consulted to see which trace to load and the "To Trace" setting to see where to load it. Trace data is always loaded with the specified trace set to View, so that the data is visible and not updating(so as not to erase the recalled data). If the file is an "all trace" file, all traces are loaded with the saved data(to the original trace the data was saved from) and set to View. Traces whose data is not loaded are restored to the update state that existed when they were saved.

In every other way a Trace load is identical to a State load. See section “File Open Dialog and Menu” on page 150

for details.

Key Path

Remote Command

Example

Recall, Trace, Open…

:MMEMory:LOAD:TRACe

TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6,<filename>

:MMEMory:LOAD:TRACe:REGister

TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6,<integer>

:MMEM:LOAD:TRAC TRACE2,"myState.trace" recalls the file myState.trace on the default path; if it is a "single trace" save file, that trace is loaded to trace 2, andis set to be not updating.

:MMEM:LOAD:TRAC:REG TRACE1,2 restores the trace data in register 2 to Trace 1

145

System Functions

Recall

Notes

Initial S/W Revision

Auto return to the Trace menu and the Open dialog goes away.

Advisory Event "Recalled File <file name>" after recall is complete.

Some modes and measurements do not have available all 6 traces. Phase

Noise mode command, for example, is: MMEMory:LOAD:TRACe

TRACE1|TRACE2|TRACE3,<filename>

The load trace command actually performs a load state, which in the Swept

SA measurement includes the trace data. However it looks in the recalled state file to see how it was flagged at save time. The possibilities are:

If the trace file was saved using one of the TRACE# enums, it is flagged as a single trace save file. The trace that was flagged as the one that was saved, is loaded to the trace specified. The trace is loaded with update off and display on, and none of the other traces are loaded.

If the trace file was saved using one the ALL enum, it is flagged as an "all traces" file. And all traces will be loaded. All of the traces are loaded with

Update=Off to keep them from updating, regardless of the setting of "Recall

State w/Trace Update".

Prior to A.02.00

Data (Import)

Importing a data file loads data that was previously saved from the current measurement or from other measurements and/or modes that produce the same type of data. The Import Menu only contains Data

Types that are supported by the current measurement.

Since the commonly exported data files are in .csv format, the data can be edited by the user prior to importing. This allows you to export a data file, manipulate the data in Excel (the most common PC

Application for manipulating .csv files) and then import it.

Importing Data loads measurement data from the specified file into the specified or default destination, depending on the data type selected. Selecting an Import Data menu key will not actually cause the importing to occur, since the analyzer still needs to know from where to get the data. Pressing the Open key in this menu brings up the Open dialog and Open menu that provides you with the options from where to recall the data. Once a filename has been selected or entered in the Open menu, the recall occurs as soon as the Open key is pressed.

Key Path

Mode

Notes

Dependencies

Recall

All

The menu is built from whatever data types are available for the mode. Some keys will be missing completely, so the key locations in the sub-menu will vary.

No SCPI command directly controls the Data Type that this key controls. The

Data Type is included in the MMEM:LOAD commands.

If a file type is not used by a certain measurement, it is grayed out for that measurement. The key for a file type will not show at all if there are no measurements in the Mode that support it.

146

System Functions

Recall

Preset

Readback

Initial S/W Revision

Is not affected by Preset or shutdown, but is reset during Restore Mode

Defaults

The data type that is currently selected

Prior to A.02.00

Trace

This key selects Trace as the data type to be imported. When pressed a second time, it brings up the

Trace Menu, which lets you select the Trace into which the data will be imported.

This key is grayed out when measurements are running that do not support trace importing.

For Vector Signal Analyzer Mode:

The trace data is loaded into the selected data register. Trace data registers are temporary storage places for trace data. They allow you to view past results next to current measurement results, and are also used in some functions like user defined filters. They are measurement global, so you can import data into a register while in the Digital Demod measurement and view it later while in the Vector measurement.

Data registers are cleared when the measurement application is terminated, but not when you change

Modes and return.

If the recalled file was saved with header information, the trace will initially be displayed with the same formatting and scaling as it had when it was saved. If headers are not saved, the scaling and format are set to defaults when the trace is recalled.

The following trace data formats may be imported:

Text and comma-separated variable (CSV)

Text

SDF.

Option 200 also allows import of these additional formats:

Matlab 4

Matlab 5

Matlab HDF5

N5110A compatible binary

Remote Command

Example

Key Path

:MMEMory:LOAD:TRACe:DATA

TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6,<filename>

:MMEM:LOAD:TRAC DATA TRACE2,"myTrace2.csv" imports the 2nd trace from the file myTrace2.csv in the current path. The default path is My

Documents\SA\data\traces

Recall, Data

147

System Functions

Recall

Notes

Dependencies

Couplings

Readback

For VSA Mode: The Open

...dialog box has the following filter options when you are recalling trace data:

• CSV (Comma delimited) (*.csv)

• SDF (Fast) (*.sdf;*.dat)

• Text (Tab delimited) (*.txt)

• MAT-File (*.mat)

• MAT-File (Version 4) (*.mat)

• MAT-File (HDF5) (*.mat;*.hdf;*.h5)

• N5110A Waveform (*.bin)

The file format recalled depends on selection.

Trace data is not available from all Measurements. When unavailable, the key will be grayed out. The key will not show if no measurements in the Mode support it.

If any error occurs while trying to load a file manually (as opposed to during remote operation), the analyzer returns to the Import Data menu and the File

Open dialog goes away.

When a trace is imported, Trace Update is always turned OFF for that trace and Trace Display is always turned ON.

Other than VSA: 1|2|3|4|5|6

VSA: Data 1|Data 2|Data 3|Data 4|Data 5|Data 6

Prior to A.02.00

Initial S/W Revision

Trace 1, 2, 3, 4, 5, 6

Enables you to select which Trace to import the data into either 1, 2, 3, 4, 5 or 6. The default is the currently selected trace, which was selected in this menu or in the Trace/Det, Export Data, Recall Trace, or Save Trace menus. The exception is, if you have chosen All then it remains chosen until you specifically change it to a single trace.

Once selected, the key returns back to the Import Data menu and the selected Trace number is annotated on the key. Now you have selected exactly what needs to be imported. To trigger an import of the selected trace, you must select the Open key in the Import Data menu.

An example of using this menu is: If you select 4 and continue to the File Open dialog, then import Trace

4 from the file selected or entered in File Name option in the File Open dialog.

Key Path

Initial S/W Revision

Recall, Data, Trace

Prior to A.02.00

Display in Selected Trace

In Vector Signal Analyzer Mode, data registers are used as temporary storage places for trace data.

A register may be displayed in any trace. If the Display in Selected Trace key is set to "Yes" then the data

148

System Functions

Recall

register into which the file is recalled is then assigned to the currently selected trace.

Example

Key Path

Mode

Initial S/W Revision

MMEM:LOAD:TRAC:DATA D1,"TRC1.TXT",TXT

This command explicitly puts the data in the specified trace.

Recall, Data (Import), Trace (to)

VSA

Prior to A.02.00

Capture Buffer

Capture Buffer functionality is not available for all measurements. The captured data is raw data

(unprocessed).

Example

Key Path

Mode

Dependencies

Initial S/W Revision

MMEM:LOAD:CAPT "MyCaptureData.bin"

This loads the file of capture data (on the default file directory path) into the instrument.

Recall, Data

WCDMA

Capture buffer data is not available from all Measurements. When unavailable, the key will be grayed out. The key will not show if no measurements in the Mode support it.

Prior to A.02.00

Zone map

A map file contains zone definitions that help simplify making measurements of frequently used signals.

The OFDMA frame structure can contain multiple-zone definitions for the uplink and downlink subframes and multiple data burst allocations. You can recall map files in which you have saved complicated OFDMA frame analysis zone definitions. This can save you time and ensure the accuracy of repeat measurements. Map files are also useful for recreating measurement settings so they can be used by other users.

Example

Key Path

Mode

Dependencies

Initial S/W Revision

MMEM:LOAD:ZMAP "MyZonemapFile.omf"

This loads the file of zone map data (on the default file directory path) into the custom map.

Recall, Data

OFDMA WIMAX

Zone map data is not available from all Measurements.When unavailable, the key will be grayed out. The key will not show if no measurements in the Mode support it.

Prior to A.02.00

149

System Functions

Recall

Recorded Data

This allows you to recall previously saved, recorded data for analysis.

This feature is only available with 89601X VSA Option 200 and Option G01.

Example

Key Path

Mode

Notes

Initial S/W Revision

MMEM:LOAD:REC "MyRecording.sdf"

Recall, Data (Import)

VSA

Available file types are:

• CSV (Comma delimited) (*.csv)

• MAT-File (*.mat)

• MAT-File (Version 4) (*.mat)

• MAT-File (HDF5) (*.mat;*.hdf;*.h5)

• N5110A Waveform (*.bin)

• SDF (Fast) (*.sdf;*.dat)

• SDF (Export) (*.sdf;*.dat)

• Text (Tab delimited) (*.txt)

Prior to A.02.00

Open…

Accesses the standard Windows File Open dialog and the File Open key menu. When you navigate to this selection, you have already determinedyou are recalling a specific Data Type and now you want to specify which file to open.

When you first enter this dialog, the path in the Look In: field depends on which import data type you selected.

The only files that are visible are those specific to the file type being recalled.

Key Path

Notes

Initial S/W Revision

Recall, Data

The key location is mode-dependent and will vary.

Brings up Open dialog for recalling a <mode specific> Save Type

Prior to A.02.00

Open

The import starts by checking for errors. Then the import can start. For all data types, the actual import starts by aborting the currently running measurement. Then the import does data type specific behavior:

File Open Dialog and Menu

The File Open is a standard Windows dialog and has a File Open key menu. Each key in this menu

150

System Functions

Recall

corresponds to the selectable items in the File Open dialog box. The menu keys can be used for easy navigation between the selections within the dialog or the standard

Tab

and

Arrow

keys can be used for dialog navigation. When you navigate to this selection, you have already limited the file recall type and now you want to specify which file to open.

Initial S/W Revision: Prior to A.02.00

Open

This selection and the

Enter

key, when a filename has been selected or specified,cause the load to occur.

Open

loads the specified or selected file to the previously selected recall type of either State or a specific import data type.

Notes

Initial S/W Revision

Advisory Event "File <file name> recalled" after recall is complete.

Prior to A.02.00

File/Folder List

This menu key navigates to the center of the dialog that contains the list of files and folders. Once hereyoucan get information about the file.

Key Path

Notes

Recall, <various>, Open…

Pressing this key navigates you to the files and folders list in the center of the dialog.

Prior to A.02.00

Initial S/W Revision

Sort

Accesses a menu that enables you to sort the files within the File Open dialog. Only one sorting type can be selected at a time and the sorting happens immediately.

Key Path

Notes

Initial S/W Revision

Recall, <various>, Open…

No SCPI command directly controls the sorting.

Prior to A.02.00

By Date

Accesses a menu that enables you to sort the list of files within the scope of the File Open dialog in ascending or descending data order. The date is the last data modified.

Key Path

Notes

Initial S/W Revision

Recall, <various>, Open…, Sort

Files in the File Open dialog are sorted immediately in the selected order

Prior to A.02.00

151

System Functions

Recall

By Name

Accesses a menu that enables you to sort the list of files within the scope of the File Open dialog in ascending or descending order based on the filename.

Key Path

Notes

Initial S/W Revision

Recall, <various>, Open…, Sort

Files in the File Open dialog are sorted immediately in the selected order

Prior to A.02.00

By Extension

Accesses a menu that enables you to sort the list of files within the scope of the File Open dialog in ascending or descending order based on the file extension for each file.

Key Path

Notes

Initial S/W Revision

Recall, <various>, Open…, Sort

Files in the File Open dialog are sorted immediately in the selected order

Prior to A.02.00

By Size

Accesses a menu that enables you to sort the list of files within the scope of the File Open dialog in ascending or descending order based on file size.

Key Path

Notes

Recall, <various>, Open…, Sort

Files in File Open dialog are sorted immediately in the selected order

Initial S/W Revision Prior to A.02.00

Ascending

This causes the display of the file list to be sorted, according to the sort criteria, in ascending order.

Key Path

Notes

Initial S/W Revision

Recall, <various>, Open…, Sort

Files in File Open dialog are sorted immediately in the selected order

Prior to A.02.00

Descending

This causes the display of the file list to be sorted, according to the sort criteria, in descending order.

Key Path

Notes

Initial S/W Revision

Recall, <various>, Open…, Sort

Files in File Open dialog are sorted immediately in the selected order

Prior to A.02.00

152

System Functions

Recall

Files Of Type

This menu key corresponds to the Files Of Type selection in the dialog. It follows the standard Windows supported Files Of Type behavior. It shows the current file suffix that corresponds to the type of file the user has selected to save. Ifyou navigated here from recalling State, "State File (*.state)" is in the dialog selection and is the only type available in the pull down menu. If you navigated here from recalling

Trace, "Trace+State File (*.trace)" is in the dialog selection and is the only type available under the pull down menu.

If younavigated here from importing a data file, the data types available will be dependent on the current measurement and the selection you made under “Import Data”. For example:

Amplitude Corrections: pull down menu shows

Amplitude Corrections (*.csv)

Legacy Cable Corrections (*.cbl)

Legacy User Corrections (*.amp)

Legacy Other Corrections (*.oth)

Legacy Antenna Corrections (*.ant)

Limit: pull down menu shows

Limit Data (*.csv)

Legacy Limit Data (*.lim)

Trace: pull down menu shows

Trace Data (*.csv)

Key Path

Notes

Initial S/W Revision

Recall, <various>, Open…

Pressing this key causes the pull down menu to list all possible file types available in this context.

Prior to A.02.00

Up One Level

This menu key corresponds to the icon of a folder with the up arrow that is in the tool bar of the dialog. It follows the standard Windows supported Up One Level behavior. When pressed, it directs the file and folder list to navigate up one level in the directory structure.

Key Path

Notes

Initial S/W Revision

Recall, <various>, Open…

When pressed, the file and folder list is directed up one level of folders and the new list of files and folders is displayed.

Prior to A.02.00

153

System Functions

Recall

Cancel

Cancels the current File Open request. It follows the standard Windows supported Cancel behavior.

Key Path

Notes

Initial S/W Revision

Recall, <various>, Open…

Pressing this key causes the Open dialog to go away and auto return.

Prior to A.02.00

154

System Functions

Save

Save

Accesses a menu that provides the save type options. The Save Type options are State, Trace, Data, or a Screen Image depending on the active mode.

Key Path

Mode

Notes

Initial S/W Revision

Save

All

No remote command for this key specifically.

Prior to A.02.00

State

Selects State as the save type and accesses a menu that provides the options of where to save. You can save either to a register or a file. This menu key will not actually cause the save until the location is chosen.

Saving the state is the only way to save this exact measurement context for the current active mode. The entire state of the active mode is saved in a way that when a recall is requested, the mode will return to as close as possible the context in which the save occurred. This includes all settings and data for only the current active mode.

It should be noted that the Input/Output settings will be saved when saving State, since these settings plus the state of the mode best characterize the current context of the mode, but the mode independent

System settings will not be saved.

For rapid saving, the State menu lists registers to save to, or you can select a file to save to. Once they select he destination of the save in the State menu, the save will occur.

Key Path

Mode

Example

Save

All

MMEM:STOR:STATe "MyStateFile.state"

This stores the current instrument state data in the file MyStateFile.state in the default directory.

See

“Save” on page 161

.

Prior to A.02.00

Notes

Initial S/W Revision

Register 1 thru Register 6

Selecting any one of these register menu keys causes the State of the currently active mode to be saved to the specified Register. The registers are provided for rapid saving and recalling, since you do not need to specify a filename or navigate to a file. Each of the register menu keys annotates whether it is empty or at what date and time it was last modified.

These 6 registers are all that is available from the front panel for all modes in the instrument. There are not 6 registers available for each mode. From remote, 127 Registers are available. Registers are files that

155

Key Path

Mode

Example

Readback

Initial S/W Revision

Key Path

Mode

Example

Readback

Initial S/W Revision

Key Path

Mode

Example

Readback

Initial S/W Revision

Key Path

Mode

Example

Readback

Initial S/W Revision

System Functions

Save

are visible to the user in the My Documents\System folder.

Key Path

Mode

Example

Readback

Initial S/W Revision

Save, State

All

*SAV 1

Date and time with seconds resolution are displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Save, State

All

*SAV 2

Date and time with seconds resolution are displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Save, State

All

*SAV 3

Date and time with seconds resolution are displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Save, State

All

*SAV 4

Date and time with seconds resolution are displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Save, State

All

*SAV 5

Date and time with seconds resolution are displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

156

System Functions

Save

Key Path

Mode

Example

Readback

Save, State

All

*SAV 6

Date and time with seconds resolution are displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Initial S/W Revision

To File . . .

Accesses a menu that enables you to select the location for saving the State. This menu is similar to a standard Windows® Save As dialog.

The default path for all State Files is:

My Documents\<mode name>\state where <mode name> is the parameter used to select the mode with the INST:SEL command (for example, SA for the Spectrum Analyzer). This path is the Save In: path in the Save As dialog for all

State Files when they first enter this dialog.

Key Path

Mode

Notes

Initial S/W Revision

Save, State

All

Brings up Save As dialog for saving a State Save Type

Prior to A.02.00

Save As . . .

Accesses a menu that enables you to select the location where you can save the State. This menu is a standard Windows® dialog with Save As menu keys. The "File Name" field in the Save As dialog is initially loaded with an automatically generated filename specific to the appropriate Save Type. The automatically generated filename is guaranteed not to conflict with any filename currently in the directory. You may replace or modify this filename using the File Name softkey. See the Quick Save key documentation for more on the automatic file naming algorithm.

The default path for all State Files is:

My Documents\<mode name>\state where <mode name> is the parameter used to select the mode with the INST:SEL command (for example, SA for the Spectrum Analyzer).

When you first enter this dialog, the path in the Save In: field depends on the data type. The only files that are visible are the *.state files and the Save As type is *.state, since .state is the file suffix for the

State Save Type.

Key Path

Mode

Save, State

All

157

System Functions

Save

Notes

Initial S/W Revision

Brings up Save As dialog for saving a State Save Type

Prior to A.02.00

Save

Saves all of the State of the currently active mode plus the system level Input/Output settings to the specified file.

While the save is being performed, the floppy icon shows up in the settings bar near the

Continuous/Single sweep icon. After the save completes, the Advisory Event "File <register number> saved" is displayed.

Key Path

Mode

Remote Command

Example

Notes

Save, State, To File…

All

:MMEMory:STORe:STATe <filename>

:MMEM:STOR:STAT "myState.state" saves the file myState.state on the default path

If the file already exists, the file will be overwritten. Using the C: drive is strongly discouraged, since it runs the risk of being overwritten during an instrument software upgrade. Both single and double quotes are supported for any filename parameter over remote.

Auto return to the State menu and the Save As dialog goes away.

Backwards Compatibility SCPI

Initial S/W Revision

For a backwards compatibility only, the following parameters syntax is supported:

:MMEMory:STORe:STATe 1,<filename>

The "1" is just ignored.

The command is sequential.

Prior to A.02.00

Trace (+State)

Selects a state file which includes trace data for recalling as the save type and accesses a menu that enables you to select which trace to save. You can save to either a register or a file. Not all modes support saving trace data with the state, and for modes that do, not all measurements do. This key is grayed out for measurements that do not support trace saves. It is blanked for modes that do not support trace saves. Saving Trace is identical to saving State except a .trace extension is used on the file instead of .state, and internal flags are set in the file indicating which trace was saved. You may also select to save ALL traces.

This key will not actually cause the save, since the save feature still needs to know which trace to save and where to save it. Pressing this key accesses the Save Trace menu that provides the user with these options.

For rapid saving, the Trace menu lists registers to save to, or you can select a file to save to. Once you

158

System Functions

Save

pick the destination of the save in the Trace menu, the save will occur.

Key Path

Mode

Example

Notes

Initial S/W Revision

Save

SA

MMEM:STOR:STATe TRACE2,"MyTraceFile.trace"

This stores trace 2 data in the file MyTraceFile.trace in the default directory.

:MMEM:STOR:TRAC:REG TRACE1,2 stores trace 1 data in trace register 2

:MMEM:STOR:TRAC:REG ALL,3 saves the data for all 6 traces in trace register 3

See

“Save” on page 161

.

Prior to A.02.00

Register 1 thru Register 5

Selecting any one of these register menu keys causes the Trace(s) specified under From Trace, along with the state of the currently active mode, to be saved to the specified Trace Register. The registers are provided for rapid saving and recalling, since you do not need to specify a filename or navigate to a file.

Each of the register menu keys annotates whether it is empty or at what date and time it was last modified.

These 5 trace registers are all that is available for all modes in the instrument. At present, only the Swept

SA measurement of the Spectrum Analyzer mode supports saving to Trace+State files. Registers are files that are visible to the user in the My Documents\System folder.

Key Path

Mode

Readback

Initial S/W Revision

Save, Trace

SA

Date and time with seconds resolution are displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Key Path

Mode

Readback

Initial S/W Revision

Key Path

Mode

Readback

Save, Trace

SA

Date and time with seconds resolution are displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Save, Trace

SA

Date and time with seconds resolution are displayed on the key, or "(empty)" if no prior save operation performed to this register.

159

System Functions

Save

Initial S/W Revision

Key Path

Mode

Readback

Initial S/W Revision

Prior to A.02.00

Save, Trace

SA

Date and time with seconds resolution are displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Key Path

Mode

Readback

Save, Trace

SA

Date and time with seconds resolution are displayed on the key, or "(empty)" if no prior save operation performed to this register.

Prior to A.02.00

Initial S/W Revision

From Trace

Accesses a menu that enables you to select the trace to be saved. Once a trace is selected, the key returns to the Save Trace menu and the selected trace number is annotated on the key. The default is the currently selected trace, selected in this menu or in the Trace/Det, Export Data, Import Data or Recall

Trace menus, except if you have chosen All then it remains chosen until you specifically change it to a single trace. To save the Trace you must select

Save As

.

These keys let you pick which trace to save. Now you have selected exactly what needs to be saved. To trigger a save of the selected Trace, you must select the

Save As

key in the Save Trace menu.

Key Path

Mode

Initial S/W Revision

Save, Trace + State

SA

Prior to A.02.00

Save As . . .

This menu lets you select the location where you can save the Trace. It is a standard Windows® dialog with Save As menu keys.

The "File Name" field in the Save As dialog is initially loaded with an automatically generated filename specific to the appropriate Save Type. The automatically generated filename is guaranteed not to conflict with any filename currently in the directory. You may replace or modify this filename using the File

Name softkey. See the Quick Save key documentation for more on the automatic file naming algorithm.

The default path for all State Files including .trace files is:

My Documents\<mode name>\state where <mode name> is the parameter used to select the mode with the INST:SEL command (for example, SA for the Spectrum Analyzer).

When you first enter this dialog, the path in the Save In: field depends on the data type. The only files

160

System Functions

Save

that are visible are the *.trace files and the Save As type is *.trace, since .trace is the file suffix for the

Trace Save Type.

Key Path

Mode

Notes

Initial S/W Revision

Save, Trace (+State)

SA

Brings up the Save As dialog for saving a Trace Save Type

Prior to A.02.00

Save

This key initiates the save of the .trace file. All of the State of the currently active mode plus the system level Input/Output settings are saved to the specified file as well as all of the trace data, including internal flags set in the file indicating which trace is to be saved.

While the save is being performed, the floppy icon shows up in the settings bar near the

Continuous/Single sweep icon. After the save completes, the Advisory Event "File <register number> saved" is displayed.

Key Path

Mode

Remote Command

Example

Notes

Save, Trace, Save As…

SA

:MMEMory:STORe:TRACe

TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6|ALL,<filename

>

:MMEMory:STORe:TRACe:REGister

TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6|ALL,<integer>

:MMEM:STOR:TRAC TRACE1,“myState.trace” saves the file myState.trace on the default path and flags it as a “single trace” file with Trace 1 as the single trace (even though all of the traces are in fact stored).

:MMEM:STOR:TRAC ALL,“myState.trace” saves the file myState.trace on the default path and flags it as an “all traces” file

:MMEM:STOR:TRAC:REG TRACE1,2 stores trace 1 data in trace register 2

Some modes and measurements do not have available all 6 traces. The Phase

Noise mode command, for example, is: MMEMory:STORe:TRACe

TRACE1|TRACE2|TRACE3|ALL,<filename>

This command actually performs a save state, which in the Swept SA measurement includes the trace data. However it flags it (in the file) as a “save trace” file of the specified trace (or all traces).

The range for the register parameter is 1–5

If the file already exists, the file will be overwritten. Using the C: drive is strongly discouraged, since it runs the risk of being overwritten during a instrument software upgrade. Both single and double quotes are supported for any filename parameter over remote.

Auto return to the State menu and the Save As dialog goes away.

161

System Functions

Save

Initial S/W Revision Prior to A.02.00

Data (Export)

Exporting a data file stores data from the current measurement to mass storage files. The Export Menu only contains data types that are supported by the current measurement.

Since the commonly exported data files are in .csv format, the data can be edited by you prior to importing. This allows youto export a data file, manipulate the data in Excel (the most common PC

Application for manipulating .csv files) and then import it.

Selecting an Export Data menu key will not actually cause the exporting to occur, since the analyzer still needs to know where you wish to save the data. Pressing the Save As key in this menu brings up the Save

As dialog and Save As menu that allows you to specify the destination file and directory. Once a filename has been selected or entered in the Open menu, the export will occur as soon as the Save key is pressed.

Key Path

Mode

Notes

Dependencies

Preset

Save

All

The menu is built from whatever data types are available for the mode. So the key locations in the sub menu will vary.

No SCPI command directly controls the Data Type that this key controls. The

Data Type is included in the MMEM:STORe commands.

If a file type is not used by a certain measurement, that type is grayed out for that measurement. The key for a file type will not show at all if there are no measurements in the Mode that support it.

Is not affected by a Preset or shutdown, but is reset during Restore Mode

Defaults

The data type that is currently selected

Prior to A.02.00

Readback

Initial S/W Revision

Trace

Enables you to select Traces as the data type to be exported. Pressing this key when it is already selected brings up the Trace menu, which allows you to select which Trace to save.

The trace file contains “meta” data, which describes the current state of the analyzer.

Key Path

Remote Command

Save, Data

:MMEMory:STORe:TRACe:DATA

TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6|ALL,<filename

>

162

System Functions

Save

Example

Notes

Dependencies

Readback

:MMEM:STOR:TRAC:DATA TRACE2,"myTrace2.csv" exports the 2nd trace to the file myTrace2.csv in the current path. The default path is My

Documents\SA\data\traces

VSA Example: MMEM:STOR:TRAC:DATA TRACE1,"Trc1.txt",TXT,ON

If the save is initiated via SCPI, and the file already exists, the file will be overwritten.

Using the C: drive is strongly discouraged, since it runs the risk of being overwritten during an instrument software upgrade.

Both single and double quotes are supported for any filename parameter over

SCPI.

Trace data is not available from all Measurements. When unavailable, the key will be grayed out. The key will not show if no measurements in the Mode support it.

Swept SA: 1|2|3|4|5|6|ALL

Analog Demod Mode: RF Spectrum|Demod|Demod Ave|Demod Max|Demod

Min|AF Spectrum

Vector Signal Analyzer: Trace 1|Trace 2 with header|Trace 2|Trace 2 with header|Trace 3|Trace 3 with header|Trace 4|Trace 4 with header|Trace 5|Trace

5 with header|Trace 6|Trace 6 with header|

Prior to A.02.00

Initial S/W Revision

Trace selection

Enables you to select which Trace to save. The traces may have names, or they may be labeled 1, 2, 3, 4,

5, or 6, depending on the current mode. Once selected, the key returns back to the Export Data menu and the selected trace name/number is annotated on the key. The default is the currently selected trace, selected in this menu or in the Trace/Det, Import Data, Recall Trace or Save Trace menus. The exception is, if you have chosen All then it remains chosen until you specifically change it to a single trace.

To trigger a save of the selected trace, you must select the Save As key in the Export Data menu.

Some measurements have an "ALL" selection. This saves all six traces in one .csv file with the x-axis data in the first column and the individual trace data in succeeding columns. The header data and x-axis data in this file reflect the current settings of the measurement. Note that any traces which are in View or

Blank may have different x-axis data than the current measurement settings, but this data will not be output to the file.

Key Path

Mode

Preset

Initial S/W Revision

Save, Data, Trace

SA|Analog Demod|VSA

The first trace key shown.

Prior to A.02.00

Include Header

The trace header information includes enough state information to display the trace data with the same

163

System Functions

Save

formatting and scaling when it is recalled. However, no other instrument state information is saved. If headers are not saved, the scaling and format are set to defaults when the trace is recalled.

Example

Key Path

Mode

Preset

Initial S/W Revision

MMEM:STOR:TRAC:DATA TRACE1,"Trc1.txt",TXT,ON

The On/Off setting is the last variable passed in the

MMEMory:STORe:TRACe:DATA command.

Save, Data, Trace

VSA

On

Prior to A.02.00

Measurement Results

Different types of results are available for each particular measurement. The results that are available are documented under the individual measurements. These measurement results are the same as the results that are returned when using the MEASure:<measurement> command (usually for sub-opcode 1).

Measurement results may not be available for all measurements.

Example

Key Path

Mode

Notes

Initial S/W Revision

MMEM:STOR:RES "MyResultsFile.xml"

This stores the measurement results data in the file MyResultsFile.xml in the default directory.

Save, Data

SA|ADEMOD|BASIC(IQ

Analyzer)|CDMA2K|GSMEDGE|PNOISE|WCDMA|WIMAXOFDMA|TDS

CDMA

The key will not show if no measurements in the Mode support it.

Prior to A.02.00

Capture Buffer

Capture Buffer functionality is not available for all measurements. The captured data is raw data

(unprocessed).

Example

Key Path

Mode

Notes

Initial S/W Revision

MMEM:STOR:CAPT "MyCaptureData.bin"

This stores the capture data in the file MyCaptureData.bin in the default directory.

Save, Data

WCDMA

The key will not show if no measurements in the Mode support it.

Prior to A.02.00

164

System Functions

Save

Zone map

A map file contains zone definitions that will help simplify making measurements of frequently used signals. The OFDMA frame structure can contain multiple-zone definitions for the uplink and downlink subframes and multiple data burst allocations. You can store map files in which you have saved complicated OFDMA frame analysis zone definitions. This can save you time and ensure the accuracy of repeated measurements. map files are also useful for recreating measurement settings so they can be used by other users.

Example

Key Path

Mode

Notes

Initial S/W Revision

MMEM:STOR:ZMAP "MyZonemapFile.omf"

This stores the zone map data in the file MyZonemapFile.omf in the default directory.

Save, Data

OFDMA WiMAX

The key will not show if no measurements in the Mode support it.

Prior to A.02.00

Recorded Data

Saving recorded data is not available for all measurements. Recorded data, and the optional header info, may be recalled later (or transferred to another instrument) for analysis.

This function is available in 89601X VSA Option 200, but not in Option 205.

Example

Key Path

Mode

Notes

Initial S/W Revision

MMEM:STOR:REC "MyRecording.sdf",SDF,ON,ON,OFF

Save, Data (Export)

VSA

Grayed out unless there is recorded data in the buffer.

Prior to A.02.00

Save As . . .

This menu lets you select the location where you can save Data Type files. It is a standard Windows® dialog with Save As menu keys. The "File Name" field in the Save As dialog is initially loaded with an automatically generated filename specific to the appropriate Save Type. The automatically generated filename is guaranteed not to conflict with any filename currently in the directory. You may replace or modify this filename using the File Name key. See the Quick Save key documentation for more on the automatic file naming algorithm.

When you first enter this dialog, the path in the Save In: field depends on the data type. The only files that are visible are the files with the corresponding data type suffix, and the Save As type lists the same suffix.

For example, if the Data Type is Amplitude Corrections, the file suffix is .csv and the *.csv files are the only visible files in the Save As dialog and .csv is the Save As Type.

The default path for saving files is:

165

System Functions

Save

For all of the Trace Data Files:

My Documents\<mode name>\data\traces

For all of the Limit Data Files:

My Documents\<mode name>\data\limits

For all of the Measurement Results Data Files:

My Documents\<mode name>\data\<measurement name>\results

For all of the Capture Buffer Data Files:

My Documents\<mode name>\data\captureBuffer

Key Path

Mode

Notes

Save, Data

All

The key location is mode-dependent and will vary.

Brings up the Save As dialog for saving a <mode specific> Save Type

Prior to A.02.00

Initial S/W Revision

Save

Saves the specified Data Type. This section describes any specific save behavior relevant to Data that is common to all modes.

When a Save of a specific Data File is requested, the specified data is saved to the specified or selected file. The save is performed immediately and does not wait until the measurement is complete.

If the file already exists, a dialog will appear that allows you to replace the existing file by selecting

OK

or you can

Cancel

the request.

While the save is being performed, the floppy icon will show up in the settings bar near the

Continuous/Single icon. After a register save completes, the corresponding register softkey annotation is updated with the date the time and an advisory message that the file was saved appears in the message bar.

Key Path

Notes

Initial S/W Revision

Save, Data, Save As…

If the file already exists, the file will be overwritten. Using the C: drive is strongly discouraged, since it runs the risk of being overwritten during a instrument software upgrade. Both single and double quotes are supported for any filename parameter over remote.

Prior to A.02.00

Screen Image

Accesses a menu of functions that enable you to specify a format and location for the saved screen image.

Pressing Screen Image brings up a menu that allows you to specify the color scheme of the Screen Image

166

System Functions

Save

(Themes) or navigate to the Save As dialog to perform the actual save.

Screen Image files contain an exact representation of the analyzer display. They cannot be loaded back onto the analyzer, but they can be loaded into your PC for use in many popular applications.

The image to be saved is actually captured when the

Save

front panel key is pressed, and kept in temporary storage to be used if the user asks for a Screen Image save. When the Screen Image softkey is pressed, a "thumbnail" of the captured image is displayed, as shown below:

When you continue on into the

Save As

menu and complete the Screen Image save, the image depicted in the thumbnail is the one that gets saved, showing the menus that were on the screen before going into the

Save

menus.

After you have completed the save, the

Quick Save

front-panel key lets you quickly repeat the last save performed, using an auto-named file, with the current screen data.

NOTE

For versions previous to A.01.55, if you initiate a screen image save by navigating through the Save menus, the image that is saved will contain the Save menu softkeys, not the menus and the active function that were on the screen when you first pressed the Save front panel key.

Key Path

Save

167

System Functions

Save

Mode

Example

All

MMEM:STOR:SCR "MyScreenFile.png"

This stores the current screen image in the file MyScreenFile.png in the default directory.

See

Prior to A.02.00

Notes

Initial S/W Revision

Themes

Accesses a menu of functions that enable you to choose the theme to be used when saving the screen image.

The Themes option is the same as the Themes option under the Display and Page Setup dialogs. It allows you to choose between themes to be used when saving the screen image.

Key Path

Remote Command

Example

Preset

Readback

Initial S/W Revision

Save, Screen Image

:MMEMory:STORe:SCReen:THEMe

TDColor|TDMonochrome|FCOLor|FMONochrome

:MMEMory:STORe:SCReen:THEMe?

:MMEM:STOR:SCR:THEM TDM

3D Color; Is not part of Preset, but is reset by Restore Misc Defaults or

Restore System Defaults All and survives subsequent running of the modes.

3D Color | 3D Mono | Flat Color | Flat Mono

Prior to A.02.00

3D Color

Selects a standard color theme with each object filled, shaded and colored as designed.

Key Path

Example

Readback

Initial S/W Revision

Save, Screen Image, Themes

MMEM:STOR:SCR:THEM TDC

3D Color

Prior to A.02.00

3D Monochrome

Selects a format that is like 3D color but shades of gray are used instead of colors.

Key Path

Example

Readback

Save, Screen Image, Themes

MMEM:STOR:SCR:THEM TDM

3D Mono

168

System Functions

Save

Initial S/W Revision Prior to A.02.00

Flat Color

Selects a format that is best when the screen is to be printed on an ink printer.

Key Path

Example

Readback

Initial S/W Revision

Save, Screen Image, Themes

MMEM:STOR:SCR:THEM FCOL

Flat Color

Prior to A.02.00

Flat Monochrome

Selects a format that is like Flat Color. But only black is used (no colors, not even gray), and no fill.

Key Path

Example

Readback

Initial S/W Revision

Save, Screen Image, Themes

MMEM:STOR:SCR:THEM FMON

Flat Mono

Prior to A.02.00

Save As…

Accesses a menu that enables you to select the location where you can save the Screen Image. This menu is a standard Windows® dialog with Save As menu keys. The Save As dialog is loaded with the file information related to the Screen Image Type. The filename is filled in using the auto file naming algorithm for the Screen Image Type and is highlighted. The only files that are visible are the *.png files and the Save As Type is *.png, since .png is the file suffix for the Screen Image Type.

The default path for Screen Images is

My Documents\<mode name>\screen.

where <mode name> is the parameter used to select the mode with the INST:SEL command (for example, SA for the Spectrum Analyzer).

This path is the Save In: path in the Save As dialog for all Screen Files when you first enter this dialog.

Key Path

Notes

Initial S/W Revision

Save, Screen Image

Brings up Save As dialog for saving a Screen Image Save Type

Prior to A.02.00

Save

Saves the screen image to the specified file using the selected theme. The image that is saved is the measurement display prior to when the Save As dialog appeared. The save is performed immediately

169

System Functions

Save

and does not wait until the measurement is complete.

Key Path

Remote Command

Example

Notes

Initial S/W Revision

Save, Screen Image, Save As…

:MMEMory:STORe:SCReen <filename>

:MMEM:STOR:SCR "myScreen.png"

If the file already exists, the file will be overwritten. Using the C: drive is strongly discouraged, since it runs the risk of being overwritten during a instrument software upgrade. Both single and double quotes are supported for any filename parameter over remote.

Auto return to the Screen Image menu and the Save As dialog goes away.

Advisory Event "File <file name> saved" after save is complete.

Prior to A.02.00

Save As . . .

Accesses a standard Windows dialog with the Save As key menu. The "File Name" field in the Save As dialog is initially loaded with an automatically generated filename specific to the appropriate Save Type.

The automatically generated filename is guaranteed not to conflict with any filename currently in the directory. You may replace or modify this filename using the File Name key. See the Quick Save key documentation for more on the automatic file naming algorithm.

The Save As dialog has the last path loaded in Save In: for this particular file type. User specified paths are remembered and persist through subsequent runs of the mode. These remembered paths are mode specific and are reset back to the default using

Restore Mode Defaults

.

Initial S/W Revision: Prior to A.02.00

Save

Performs the actual save to the specified file of the selected type. The act of saving does not affect the currently running measurement and does not require you to be in single measurement mode to request a save. It performs the save as soon as the currently running measurement is in the idle state; when the measurement completes. This ensures the State or Data that is saved includes complete data for the current settings. The save only waits for the measurement to complete when the state or data that depends on the measurement setup is being saved. The save happens immediately when exporting corrections or when saving a screen image.

If the file already exists, a dialog appears with corresponding menu keys that allowyou to replace the existing file with an

OK

or to

Cancel

the request.

While the save is being performed, the floppy icon shows up in the settings bar near the

Continuous/Single icon. After the save completes, the corresponding register menu key annotation is

170

System Functions

Save

updated with the date the time and the message "File <file name> saved" appears in the message bar.

Notes

Initial S/W Revision

If the file already exists, the File Exist dialog appears and allows you to replace it or not by selecting the Yes or No menu keys that appear with the dialog.Then the key causes an auto return and Save As dialog goes away.

Advisory Event "File <file name> saved" after save is complete.

Prior to A.02.00

File/Folder List

Enables you to navigate to the center of the dialog that contains the list of files and folders. Once here you can get information about the file.

Key Path

Notes

Initial S/W Revision

Save, <various>, Save As…

Pressing this key enables you to navigate to the files and folders list in the center of the dialog.

Prior to A.02.00

File Name

Accesses the Alpha Editor. Use the knob to choose the letter to add and the Enter front-panel key to add the letter to the file name. In addition to the list of alpha characters, this editor includes a

Space

key and a

Done

key. The

Done

key completes the filename, removes the Alpha Editor and returns back to the

File Open dialog and menu, but does not cause the save to occur. You can also use

Enter

to complete the file name entry and this will cause the save to occur.

Key Path

Notes

Initial S/W Revision

Save, <various>, Save As…

Brings up the Alpha Editor. Editor created file name is loaded in the File name field of the Save As dialog.

Prior to A.02.00

Save As Type

This key corresponds to the Save As Type selection in the dialog. It follows the standard Windows® supported Save As Type behavior. It shows the current file suffix that corresponds to the type of file you have selected to save. If you navigated here from saving State, "State File (*.state)" is in the dialog selection and is the only type available under the pull down menu. If you navigated here from saving

Trace, "Trace+State File (*.trace)" is in the dialog selection and is the only type available under the pull down menu. If you navigated here from exporting a data file, "Data File (*.csv)" is in the dialog and is available in the pull down menu. Modes can have other data file types and they would also be listed in the pull down menu.

Key Path

Notes

Save, <various>, Save As…

Pressing this key causes the pull down menu to list all possible file types available in this context. All types available are loaded in a 1-of-N menu key for easy navigation.

171

System Functions

Save

Initial S/W Revision Prior to A.02.00

Up One Level

This key corresponds to the icon of a folder with the up arrow that is in the tool bar of the dialog. It follows the standard Windows® supported Up One Level behavior. When pressed, it causes the file and folder list to navigate up one level in the directory structure.

Key Path

Notes

Save, <various>, Save As…

When pressed, the file and folder list is directed up one level of folders and the new list of files and folders is displayed

Prior to A.02.00

Initial S/W Revision

Create New Folder

This key corresponds to the icon of a folder with the "*" that is in the tool bar of the dialog. It follows the standard Windows® supported Create New Folder behavior. When pressed, a new folder is created in the current directory with the name New Folder and allows you to enter a new folder name using the

Alpha Editor.

Key Path

Notes

Save, <various>, Save As…

Creates a new folder in the current folder and lets the user fill in the folder name using the Alpha Editor.

Prior to A.02.00

Initial S/W Revision

Cancel

This key corresponds to the

Cancel

selection in the dialog. It follows the standard Windows supported

Cancel behavior. It causes the current Save As request to be cancelled.

Key Path

Notes

Save, <various>, Save As…

Pressing this key causes the Save As dialog to go away and auto return.

Initial S/W Revision Prior to A.02.00

172

System Functions

Save

Mass Storage Catalog (Remote Command Only)

Remote Command:

Notes:

Initial S/W Revision:

:MMEMory:CATalog? [<directory_name>]

The string must be a valid logical path.

Query disk usage information (drive capacity, free space available) and obtain a list of files and directories in a specified directory in the following format:

<numeric_value>,<numeric_value>,{<file_entry>}

It shall return two numeric parameters and as many strings as there are files and directories. The first parameter shall indicate the total amount of storage currently used in bytes. The second parameter shall indicate the total amount of storage available, also in bytes. The <file_entry> is a string. Each

<file_entry> shall indicate the name, type, and size of one file in the directory list:

<file_name>,<file_type>,<file_size>

As windows file system has an extension that indicates file type, <file_type> is always empty. <file_size> provides the size of the file in bytes. In case of directories, <file_entry> is surrounded by square brackets and both

<file_type> and <file_size> are empty.

Prior to A.02.00

Mass Storage Change Directory (Remote Command Only)

Remote Command:

Notes:

Initial S/W Revision:

:MMEMory:CDIRectory [<directory_name>]

:MMEMory:CDIRectory?

The string must be a valid logical path.

Changes the default directory for a mass memory file system. The

<directory_name> parameter is a string. If no parameter is specified, the directory is set to the *RST value.

At *RST, this value is set to the default user data storage area, that is defined as System.Environment.SpecialFolder.Personal.

Query returns full path of the default directory.

Prior to A.02.00

Mass Storage Copy (Remote Command Only)

Remote Command:

:MMEMory:COPY <string>,<string>[,<string>,<string>]

173

System Functions

Save

Notes: The string must be a valid logical path.

Copies an existing file to a new file or an existing directory to a new directory.

Two forms of parameters are allowed. The first form has two parameters. In this form, the first parameter specifies the source, and the second parameter specifies the destination.

The second form has four parameters. In this form, the first and third parameters specify the source. The second and fourth parameters specify the directories. The first pair of parameters specifies the source. The second pair specifies the destination. An error is generated if the source doesn’t exist or the destination file already exists.

Mass Storage Delete (Remote Command Only)

Remote Command:

Notes:

Initial S/W Revision:

:MMEMory:DELete <file_name>[,<directory_name>]

The string must be a valid logical path.

Removes a file from the specified directory. The <file_name> parameter specifies the file name to be removed.

Prior to A.02.00

Mass Storage Data (Remote Command Only)

Creates a file containing the specified data OR queries the data from an existing file.

Remote Command:

Notes:

Initial S/W Revision:

:MMEMory:DATA <file_name>, <data>

:MMEMory:DATA? <file_name>

The string must be a valid logical path.

The command form is MMEMory:DATA <file_name>,<data>. It loads

<data> into the file <file_name>. <data> is in 488.2 block format.

<file_name> is string data.

The query form is MMEMory:DATA? <file_name> with the response being the associated <data> in block format.

Prior to A.02.00

174

System Functions

Save

Mass Storage Make Directory (Remote Command Only)

Remote Command:

Notes:

Initial S/W Revision:

:MMEMory:MDIRectory <directory_name>

The string must be a valid logical path.

Creates a new directory. The <directory_name> parameter specifies the name to be created.

Prior to A.02.00

Mass Storage Move (Remote Command Only)

Remote Command:

Notes:

Initial S/W Revision:

:MMEMory:MOVE <string>,<string>[,<string>,<string>]

The string must be a valid logical path.

Moves an existing file to a new file or an existing directory to a new directory.

Two forms of parameters are allowed. The first form has two parameters. In this form, the first parameter specifies the source, and the second parameter specifies the destination.

The second form has four parameters. In this form, the first and third parameters specify the source. The second and fourth parameters specify the directories. The first pair of parameters specifies the source. The second pair specifies the destination. An error is generated if the source doesn’t exist or the destination file already exists.

Prior to A.02.00

Mass Storage Remove Directory (Remote Command Only)

Remote Command:

Notes:

Initial S/W Revision:

:MEMMory:RDIRectory <directory_name>

The string must be a valid logical path.

Removes a directory. The <directory_name> parameter specifies the directory name to be removed. All files and directories under the specified directory shall also be removed.

Prior to A.02.00

175

System Functions

System

System

Opens a menu of keys that access various configuration menus and dialogs.

Key Path

Notes

Initial S/W Revision

Front-panel key

No remote command for this key specifically.

Prior to A.02.00

Show

Accesses a menu of choices that enable you to select the information window you want to view.

Key Path

Mode

Remote Command

Example

Notes

Preset

State Saved

Range

Initial S/W Revision

System

All

:SYSTem:SHOW

OFF|ERRor|SYSTem|HARDware|LXI|HWSTatistics|ALIGNment|SO

FTware

:SYSTem:SHOW?

:SYST:SHOW SYST

This command displays (or exits) the various System information screens.

OFF

No

OFF| ERRor | SYSTem | HARDware | LXI | HWSTatistics | ALIGNment |

SOFTware

Prior to A.02.00

Errors

There are two modes for the Errors selection, History and Status.

The list of errors displayed in the Errors screen does not automatically refresh. You must press the

Refresh key or leave the screen and return to it to refresh it.

History brings up a screen displaying the event log in chronological order, with the newest event at the top. The history queue can hold up to 100 messages (if a message has a repeat count greater than 1 it only counts once against this number of 100). Note that this count bears no relation to the size of the SCPI queue. If the queue extends onto a second page, a scroll bar appears to allow scrolling with a mouse.

Time is displayed to the second.

Status brings up a screen summarizing the status conditions currently in effect. Note that the time is displayed to the second.

The fields on the Errors display are:

176

System Functions

System

Type (unlabelled) - Displays the icon identifying the event or condition as an error or warning.

ID - Displays the error number.

Message - Displays the message text.

Repeat (RPT) - This field shows the number of consecutive instances of the event, uninterrupted by other events. In other words, if an event occurs 5 times with no other intervening event, the value of repeat will be 5.

If the value of Repeat is 1 the field does not display. If the value of Repeat is >1, the time and date shown are those of the most recent occurrence. If the value of repeat reaches 999,999 it stops there.

Time - Shows the most recent time (including the date) at which the event occurred.

Key Path

Mode

Remote Command

Example

Notes

State Saved

Initial S/W Revision

System, Show

All

:SYSTem:ERRor[:NEXT]?

:SYST:ERR?

The return string has the format:

“<Error Number>,<Error>”

Where <Error Number> and <Error> are defined in the Master Error

Messages document.

No

Prior to A.02.00

Next Page

Next Page and Previous Page menu keys move you between pages of the log, if it fills more than one page. These keys are grayed out in some cases:

If on the last page of the log, the Next Page key is grayed out

If on the first page of the log, the Previous Page key is grayed out.

If there is only one page, both keys are grayed out.

Key Path

Initial S/W Revision

Previous Page

See “Next Page” on page 177 .

System, Show, Errors

Prior to A.02.00

Key Path

Initial S/W Revision

System, Show, Errors

Prior to A.02.00

177

System Functions

System

History

The History and Status keys select the Errors view. The Status key has a second line which shows a number in [square brackets]. This is the number of currently open status items.

Key Path

Initial S/W Revision

System, Show, Errors

Prior to A.02.00

Status

See

“History” on page 178

Verbose SCPI On/Off

This is a capability that will allow the SCPI data stream to be displayed when a SCPI error is detected, showing the characters which stimulated the error and several of the characters preceding the error.

Key Path

Mode

Remote Command

Example

Preset

Preset

State Saved

Range

Initial S/W Revision

System, Show, Errors

All

:SYSTem:ERRor:VERBose OFF|ON|0|1

:SYSTem:ERRor:VERBose?

:SYST:ERR:VERB ON

This is unaffected by Preset but is set to OFF on a “Restore System

Defaults->Misc”

OFF

No

On | Off

Prior to A.02.00

Refresh

When pressed, refreshes the Show Errors display.

Key Path

Initial S/W Revision

System, Show, Errors

Prior to A.02.00

Clear Error Queue

This clears all errors in all error queues.

Note the following:

Clear Error Queue

does not affect the current status conditions.

Mode Preset

does not clear the error queue.

Restore System Defaults will clear all error queues.

178

System Functions

System

*CLS only clears the queue if it is sent remotely and *RST does not affect any error queue.

Switching modes does not affect any error queues.

Key Path

Initial S/W Revision

System, Show, Errors

Prior to A.02.00

System

The System screen is formatted into three groupings: product descriptive information, options tied to the hardware, and software products:

<Product Name> <Product Description>

Product Number: N9020A

Serial Number: US46220924

Firmware Revision: A.01.01

Computer Name: <hostname>

Host ID: N9020A,US44220924

N9020A-503

N9020A-PFR

N9020A-P03

N9060A-2FP

N9073A-1FP

N9073A-2FP

Frequency Range to 3.6 GHz

Precison Frequency Reference

Preamp 3.6 GHz

Spectrum Analysis Measurement Suite 1.0.0.0

WCDMA

WCDMA with HSDPA

1.0.0.0

1.0.0.0

The Previous Page is grayed-out if the first page of information is presently displayed. The Next Page menu key is grayed-out if the last page is information is presently displayed.

Key Path

Mode

Example

Initial S/W Revision

System, Show

All

SYST:SHOW SYST

Prior to A.02.00

Hardware

The show hardware screen is used to view details of the installed hardware. This information can be used to determine versions of hardware assemblies and field programmable devices, in the advent of future upgrades or potential repair needs.

The screen is formatted into two groupings: product descriptive information and hardware information.

The hardware information is listed in a table format:

179

System Functions

System

The Previous Page is grayed-out if the first page of information is presently displayed. The Next Page menu key is grayed-out if the last page is information is presently displayed.

Key Path

Mode

Example

Initial S/W Revision

System, Show

All

SYST:SHOW HARD

Prior to A.02.00

LXI

This key shows you the product number, serial number, firmware revision, computer name, IP address,

Host ID, LXI Class, LXI Version, MAC Address, and the Auto-MDIX Capability.

Key Path

Initial S/W Revision

System, Show

Prior to A.02.00

LXI Event Log

The event log records all of the LXI LAN event activity. As LXI LAN events are sent or received, the

180

System Functions

System

activity is noted in the Event Log with an IEEE 1588 timestamp. When the event log is selected, the current contents of the event log are displayed in the system information screen.

The fields recorded in the Event Log are:

• The date the event occurred (GMT)

• The time the event occurred (GMT)

• The type of event: LAN Input, LAN Output, Status, Alarm, Trigger Alarm, Trigger LAN

• The name of the event

• The edge associated with the event

• The event’s identifier: This is the string that appears on the LAN.

• The source event: This is only valid for LAN Output, Trigger LAN, and Trigger Alarm event types.

• The source address: This is only valid for LAN Input event types. It is the address from which the message originated.

• The destination address: This is only valid for LAN Output event types. It is the address (or addresses) that the message will be sent to. For UDP messages, this field reads “ALL.”

Key Path

Initial S/W Revision

Next Page

See

“Next Page” on page 177

.

Key Path

Initial S/W Revision

Previous Page

See

“Next Page” on page 177

System, Show, LXI

Prior to A.02.00

System, Show, Errors

Prior to A.02.00

Key Path

Initial S/W Revision

System, Show, Errors

Prior to A.02.00

Circular

Sets the behavior for entries that occur while the LXI Event Log is full.

• If Circular is set to 1, incoming events overwrite the oldest events in the log.

• If Circular is set to 0, incoming events are discarded.

Key Path

System, Show, LXI, LXI Event Log

181

System Functions

System

Remote Command

Example

Preset

Preset

State Saved

Range

Initial S/W Revision

Clear

Clears the event log of all entries.

:LXI:EVENt:LOG:CIRCular[:ENABle] ON|OFF|1|0

:LXI:EVENt:LOG:CIRCular[:ENABle]?

:LXI:EVEN:LOG:CIRC 1

ON

Not affected by a Preset. The default value of "ON" can be restored by pressing

System, Restore Defaults, Misc.

Saved in instrument state.

OFF|ON|0|1

Prior to A.02.00

Key Path

Remote Command

Example

Initial S/W Revision

System, Show, LXI, LXI Event Log

:LXI:EVENt:LOG:CLEar

:LXI:EVEN:LOG:CLE

Prior to A.02.00

Size

Sets the maximum number of entries the LXI Event Log can hold.

Key Path

Remote Command

Example

Preset

Preset

State Saved

Range

Initial S/W Revision

System, Show, LXI, LXI Event Log

:LXI:EVENt:LOG:SIZE <size>

:LXI:EVENt:LOG:SIZE?

:LXI:EVEN:LOG:SIZE 256

64

Not affected by a Preset. The default value of "64" can be restored by pressing

System, Restore Defaults, Misc

.

Saved in instrument state.

>= 0

Prior to A.02.00

Enabled

Enables and disables the logging of LXI Events.

Key Path

System, Show, LXI, LXI Event Log

182

System Functions

System

Remote Command

Example

Preset

Preset

State Saved

Range

Initial S/W Revision

:LXI:EVENt:LOG:ENABle ON|OFF|1|0

:LXI:EVENt:LOG:ENABle?

:LXI:EVEN:LOG:ENAB ON

ON

Not affected by a Preset. The default value of "ON" can be restored by pressing

System, Restore Defaults, Misc.

Saved in instrument state.

ON|OFF|0|1

Prior to A.02.00

Count (Remote Command Only)

Returns the number of entries currently in the LXI Event Log.

Remote Command:

Example:

Range:

Initial S/W Revision:

:LXI:EVENt:LOG:COUNt?

:LXI:EVEN:LOG:COUN?

0 – Size

Prior to A.02.00

Next Entry (Remote Command Only)

Returns the oldest entry from the LXI Event Log and removes it from the log. If the log is empty, an empty string is returned.

Remote Command:

Example:

Initial S/W Revision:

:LXI:EVENt:LOG[:NEXT]?

:LXI:EVEN:LOG?

Prior to A.02.00

All (Remote Command Only)

Non-destructively retrieves the entire contents of the event log. Entries are returned as separate strings, surrounded by double quote marks, and separated by a comma. Fields within each entry are also comma delimited.

Remote Command:

Example:

:LXI:EVENt:LOG:ALL?

:LXI:EVEN:LOG:ALL? !Returns the entire event log contents.

An example may look like the following:

"11/12/2007,18:14:10.770385,Error,LogOverwrite,Rise,,,,","11/12/2007,18:1

4:10.592105,Status,Measuring,Rise,,,,","11/12/2007,18:14:10.597758,Status,

Measuring,Fall,,,,","11/12/2007,18:14:10.597786,Status,Sweeping,Fall,,,,","1

1/12/2007,18:14:10.599030,Status,WaitingForTrigger,Rise,,,,"

The contents of the Event Log vary, based on the operation of the instrument.

183

System Functions

System

Initial S/W Revision: Prior to A.02.00

Specific Entry (Remote Command Only)

Non-destructively retrieves a specifically indexed entry from the event log. Fields within an entry are comma delimited.

Remote Command:

Example:

:LXI:EVENt:LOG:ENTRy? <intIndex>

:LXI:EVEN:LOG:ENTR? 0 !Returns the first entry in the event log.

An example may look like the following:

"11/12/2007,18:14:10.770385,Error,LogOverwrite,Rise,,,,"

The contents of the Event Log vary, based on the operation of the instrument.

Prior to A.02.00

Initial S/W Revision:

Beginning Entry (Remote Command Only)

Sets or freezes the beginning entry of the log when in circular mode to the most recently added entry at the time of the command. This is so that the :LXI:EVENt:LOG:ENTtry? command has a reference entry for indexing individual entries in the log.

Remote Command:

Example:

Initial S/W Revision:

:LXI:EVENt:LOG:CIRCular:FBENtry

:LXI:EVEN:LOG:CIRC:FBEN

Prior to A.02.00

Power On

Enables you to select how the instrument should power on. The options are: Mode and Input/Output

Defaults, User Preset and Last State.

Key Path

Mode

Remote Command

Example

Preset

Preset

State Saved

Initial S/W Revision

System

All

:SYSTem:PON:TYPE MODE|USER|LAST|PRESet

:SYSTem:PON:TYPE?

:SYST:PON:TYPE MODE

MODE

This is unaffected by a Preset but is set to Mode on a “Restore System

Defaults->All”

No

Prior to A.02.00

184

System Functions

System

Mode and Input/Output Defaults

When the analyzer is powered on in Mode and Input/Output Defaults, it will perform a Restore Mode

Defaults to all modes in the instrument and a Restore Input/Output Defaults as well.

Key Path

Mode

Example

Readback Text

Initial S/W Revision

System, Power On

All

SYST:PON:TYPE MODE

Defaults

Prior to A.02.00

User Preset

Sets

Power On

to

User Preset

. When the analyzer is powered on in User Preset, it will User Preset each mode and switch to the power-on mode. Power On User Preset will not affect any settings beyond what a normal User Preset affects.

NOTE

An instrument could never power up for the first time in User Preset.

Key Path

Mode

Example

Readback Text

Initial S/W Revision

System, Power On

All

SYST:PON:TYPE USER

User Preset

Prior to A.02.00

Last State

Sets

Power On

to Last. When the analyzer is powered on, it will put all modes in the last state they were in prior to when the analyzer was put into Power Standby and it will wake up in the mode it was last in prior to powering off the instrument. The saving of the active mode prior to shutdown happens behind the scenes when a controlled shutdown is requested by using the front panel power

Standby

key or by using the remote command SYSTem:PDOWn. The non-active modes are saved as they are deactivated and recalled by Power On Last State.

NOTE

An instrument could never power up for the first time in Last.

If line power to the analyzer is interrupted, for example by pulling the line cord plug or by switching off power to a test rack, Power On Last State will not work properly.

185

System Functions

System

Key Path

Mode

Example

Notes

Readback Text

Initial S/W Revision

System, Power On

All

SYST:PON:TYPE LAST

Power on Last State only works if the user has done a controlled shutdown prior to powering on in Last. If a controlled shutdown is not done when in

Power On Last State, the instrument will power up in the last active mode, but it may not power up in the active mode’s last state. If an invalid mode state is detected, a Mode Preset will occur. To control the shutdown under remote control use the :SYSTem:PDOWn command.

Last State

Prior to A.02.00

Power On Application

Accesses a menu that lists the available Modes and lets you select which Mode is to be the power-on application.

This application is used for Power On Type “Mode and Input/Output Defaults” and Restore System

Defaults All.

Key Path

Mode

Remote Command

Example

Notes

Preset

Preset

State Saved

Initial S/W Revision

System, Power On

All

:SYSTem:PON:MODE

SA|BASIC|ADEMOD|NFIGURE|PNOISE|CDMA2K|TDSCDMA|VSA|VSA89

601|WCDMA|WIMAXOFDMA

:SYSTem:PON:MODE?

SYST:PON:MODE SA

The list of possible modes (and remote parameters) to choose from is dependent on which modes are installed in the instrument.

SA

This is unaffected by a Preset but is set on a “Restore System Defaults->All” to SA.

No

Prior to A.02.00

Configure Applications

The Configure Applications utility lets you do two things:

1. specify a subset of the available applications (Modes) to preload into memory at startup time

2. specify the order in which the Modes appear in the Mode menu

186

System Functions

System

There are several reasons you might want to specify a subset of the available applications (Modes) to preload:

• During runtime, if a Mode which is not preloaded is selected by the user, there will be a pause while the Application is loaded. Configure Applications lets you decide whether you want that delay at startup of the analyzer program or the first time you select the Mode.

• In addition, there are more applications available for the X-Series than can fit into Windows Virtual

Memory. The Configure Application utility allows you to choose which licensed applications to load into memory, if you have more licensed than can fit.

The Configure Applications utility can be used to select applications for preload and/or to determine how many applications can fit in memory at one time. This utility consists of a window with instructions, a set of “Select Application” checkboxes, a “fuel bar” style memory guage, and softkeys that help you set up your configuration.

For more information see:

“Preloading Applications” on page 187

“Virtual memory usage” on page 187

“Access to Configure Applications utility” on page 188

Preloading Applications

During operation of the analyzer, you select applications from the Mode menu. After startup of the analyzer program, the first time you select a particular application that application must be loaded into memory. Once loaded, the application stays loaded, so the next time you select it during a session, there is no delay. During runtime, if an application which is not yet loaded into memory is selected using the

Mode menu or sending SCPI commands, there will be a pause while the Application is loaded. During this pause a message which says “Loading application, please wait …” is displayed.

You can use the Configure Applications utility to choose applications to “preload” at startup, to eliminate the runtime delay; if you do this, the delay will instead increase the time it takes to start up the analyzer program, but for many users this is preferable to having to wait the first time they select an application.

Asking for an application to be preloaded will cause it to be loaded into the analyzer’s memory when the analyzer program starts up. Once it is loaded into memory, it cannot be unloaded without exiting and restarting the analyzer program.

Virtual memory usage

There are more applications available for the X-Series than can fit into memory at any one time, so the

Configure Applications utility includes a memory tracker that serves two purposes:

It will not let you preload more applications than will fit into memory at once.

You can determine how many of your favorite applications can reside in memory at one time.

The utility provides a graphical representation of the amount of memory (note that the memory in question here is Virtual memory and is a limitation imposed by the oprating system, not by the amount of physical memory you have in your analyzer). You select applications to preload by checking the boxes on the left. Checked applications preload at startup. The colored fuel bar indicates the total memory required when all the checked applications are loaded (either preloaded or selected during runtime).

Here is what the fuel bar colors mean:

187

System Functions

System

RED: the applications you have selected cannot all fit into the analyzer’s memory. You must deselect applications until the fuel bar turns yellow.

YELLOW: the applications you have selected can all fit into the analyzer’s memory, but there is less than 10% of the memory left, probably not enough to load any other applications, either via preload or by selecting a Mode while the analyzer is running..

GREEN: The indicator is green when <90% of the memory limit is consumed. This means the applications you have selected can all fit into the analyzer’s memory with room to spare. You will likely be able to load one or more other applications without running out of memory.

Access to Configure Applications utility

You may, at any time, manually call up the Configure Applications utility by pressing

System, Power

On, Configure Applications

, to find a configuration that works best for you, and then restart the analyzer program.

The utility may also be called if, during operation of the analyzer, you attempt to load more applications than can fit in memory at once.

A version of the utility also runs the first time you power up the analyzer after purchasing it from

Agilent. In this case the utility automatically configures preloads so that as many licensed applications as possible are preloaded while keeping the total estimated virtual memory usage below the limit. This auto-configuration only takes place at the very first run, and after analyzer software upgrades.

Key Path

Initial S/W Revision

System, Power On

A.02.00

Select All

Marks all applications in the selection list. This allows you to enable all applications licensed on the instrument for pre-loading, or is a convenience for selecting all applications in one operation and then letting you deselect individual applications.

Key Path

Initial S/W Revision

System, Power On, Configure Applications

A.02.00

Deselect All

Clears the marks from all applications in the selection list, with the exception of the Power On application. The Power On application cannot be eliminated from the pre-load list.

Key Path

Initial S/W Revision

System, Power On, Configure Applications

A.02.00

Move Up

The application list is the order in which applications appear in the Mode Menu. This key enables you to shift the selected application up in the list, thus moving the selected application earlier in the Mode

188

System Functions

System

Menu.

Key Path

Initial S/W Revision

System, Power On, Configure Applications

A.02.00

Move Down

The application list is the order in which applications appear in the Mode Menu. This key enables you to shift the selected application down in the list, thus moving the selected application later in the Mode

Menu.

Key Path

Initial S/W Revision

System, Power On, Configure Applications

A.02.00

Select/Deselect

Toggles the currently highlighted application in the list.

Key Path

Initial S/W Revision

System, Power On, Configure Applications

A.02.00

Save Changes and Exit

Applies the configuration of the applications list. The marked applications will be pre-loaded in memory the next time the instrument application is started, and the order of the applications in the list will be the order of the applications in the Mode Menu.

After saving your changes, the analyzer asks you if you would like it to restart so that your changes can take effect (see dialog box, below). If you choose not to restart, no memory will be released until the next time you shut down and restart the analyzer.

Key Path

Notes

Remote Command

Example

System, Power On, Configure Applications

The softkey will be grayed-out when the virtual memory of the selected applications exceeds 100% of the limit.

:SYSTem:PUP:PROCess

:SYST:PUP:PROC ! This is the SCPI command for restarting the analyzer.

You must Wait after this command for the instrument application to restart

189

System Functions

System

Notes

Initial S/W Revision

Modified at S/W Revision

You cannot use *WAI or *OPC? to synchronize operation after a restart. This command stops and restarts the instrument application, thus the SCPI operation is terminated and restarted. A remote program must use fixed wait time to resume sending commands to the instrument. The wait time will be dependent upon which applications are pre-loaded.

A.02.00

A.04.00

Exit Without Saving

Pressing this key will exit the Configure Applications utility without saving your changes.

Key Path

Initial S/W Revision

Modified at S/W Revision

System, Power On, Configure Applications

A.02.00

A.04.00

Configure Applications - Instrument boot-up

At start-up of the analyzer programa dialog box similar to the one under the

System, Power On,

Configure Applications

key will be displayed allowing you to choose which licensed applications are to be loaded. This dialog will only be displayed if the memory required to pre-load all of the licensed applications exceeds the Virtual Memory available.

Configure Applications - Windows desktop

The Configure Applications Utility may be run from the Windows Desktop. The utility is launched by double-clicking the icon on the desktop, which brings-up a dialog box similar to the one under the

System, Power On, Configure Applications

key, allowing you to choose which licensed applications are to be loaded whne the analyzer program starts up. This dialog box has mouse buttons on it which do the job that the softkeys normally do in the

System, Power On, Configure Applications

menu.

Configure Applications - Remote Commands

The following topics provide details on the using remote commands to configure the list of applications want to load into the instrument memory or query the Virtual Memory utilization for your applications.

“Configuration list (Remote Command Only)” on page 191

“Configuration Memory Available (Remote Command Only)” on page 191

“Configuration Memory Total (Remote command Only)” on page 191

“Configuration Memory Used (Remote Command Only)” on page 191

“Configuration Application Memory (Remote Command Only)” on page 192

190

System Functions

System

Configuration list (Remote Command Only)

This remote command is used to set or query the list of applications to be loaded in-memory.

Remote Command:

Example:

Notes:

Preset:

State Saved:

Initial S/W Revision:

:SYSTem:PON:APPLication:LLISt <string of

INSTrument:SELect names>

:SYSTem:PON:APPLication:LLISt?

:SYST:PON:APPL:LLIS "SA,BASIC,WCDMA"

<string of INSTrument:SELect names> are from the enums of the

:INSTrument:SELect command.

The order of the <INSTrument:SELect names> is the order in which the applications are loaded into memory, and the order in which they appear in the

Mode Menu.

Error –225 "Out of Memory" is reported when more applications are listed than can reside in Virtual Memory. When this occurs, the existing applications load list is unchanged.

Not affected by Preset

Not saved in state

A.02.00

Configuration Memory Available (Remote Command Only)

This remote command is used to query the amount of Virtual Memory remaining.

Remote Command:

Example:

Preset:

Initial S/W Revision:

:SYSTem:PON:APPLication:VMEMory[:AVAilable]?

:SYST:PON:APPL:VMEM?

Not affected by Preset

A.02.00

Configuration Memory Total (Remote command Only)

This remote command is used to query the limit of Virtual Memory allowed for applications.

Remote Command:

Example:

Preset:

Initial S/W Revision:

:SYSTem:PON:APPLication:VMEMory:TOTal?

:SYST:PON:APPL:VMEM:TOT?

Not affected by Preset

A.02.00

Configuration Memory Used (Remote Command Only)

This remote command is a query of the amount of Virtual Memory used by all measurement

191

System Functions

System

applications.

Remote Command:

Example:

Preset:

Initial S/W Revision:

:SYSTem:PON:APPLication:VMEMory:USED?

:SYST:PON:APPL:VMEM:USED?

Not affected by Preset

A.02.00

Configuration Application Memory (Remote Command Only)

This remote command is used to query the amount of Virtual Memory a particular application consumes.

Remote Command:

Example:

Notes:

Preset:

Initial S/W Revision:

:SYSTem:PON:APPLication:VMEMory:USED:NAME?

<INSTrument:SELect name>

:SYST:PON:APPL:VMEM:USED:NAME? CDMA2K

<INSTrument:SELect name> is from the enums of the :INSTrument:SELect command in Meas Common section 13.3

Value returned will be 0 (zero) if the name provided is invalid.

Not affected by Preset

A.01.70 or later

Restore Power On Defaults

This selection causes the Power On Type and Power On Application settings to be a reset to their default values. This level of Restore System Defaults does not affect any other system settings, mode settings and does not cause a mode switch. The Power On key, under the Restore System Defaults menu, causes the same action.

If you press any key other than OK or Enter, it is construed as a Cancel, because the only path that will actually cause the reset to be executed is through OK or Enter.

Key Path

Example

System, Power On

:SYST:DEF PON

192

System Functions

System

Initial S/W Revision Prior to A.02.00

Alignments

The Alignments Menu controls and displays the automatic alignment of the instrument, and provides the ability to restore the default alignment values.

The current setting of the alignment system is displayed in the system Settings Panel along the top of the display, including a warning icon for conditions that may cause specifications to be impacted.

Key Path

Initial S/W Revision

System

Prior to A.02.00

Auto Align

Configures the method for which the automatic background alignment is run.

Automatic background alignments are run periodically between measurement acquisitions. The instrument’s software determines when alignments are to be performed to maintain warranted operation.

The recommended setting for Auto Align is Normal.

An Auto Align execution cannot be aborted with the Cancel (ESC) key. To interrupt an Auto Align execution, select

Auto Align Off

.

Key Path

Mode

Remote Command

Example

Notes

Couplings

Preset

Preset

State Saved

Status Bits/OPC dependencies

Initial S/W Revision

System, Alignments

All

:CALibration:AUTO ON|PARTial|OFF

:CALibration:AUTO?

:CAL:AUTO ON

While Auto Align is executing, bit 0 of Status Operation register is set.

Auto Align is set to Off if Restore Align Data is invoked.

ON

This is unaffected by Preset but is set to ON upon a “Restore System

Defaults->Align”.

No

When Auto Align is executing, bit 0 in the Status Operational register is set.

Prior to A.02.00

Normal

Auto Align, Normal

turns on the automatic alignment of all measurement systems. The Auto Align,

Normal selection maintains the instrument in warranted operation across varying temperature and over

193

System Functions

System

time.

If the condition “Align Now, All required” is set, transition to Auto Align, Normal will perform the required alignments and clear the “Align Now, All required” condition and then continue with further alignments as required to maintain the instrument adequately aligned for warranted operation.

When

Auto Align

,

Normal

is selected the Auto Align Off time is set to zero.

When

Auto Align

,

Normal

is selected the Settings Panel indicates ALIGN AUTO.

Key Path

Mode

Example

Notes

Readback Text

Status Bits/OPC dependencies

System, Alignments, Auto Align

All

:CAL:AUTO ON

Alignment processing as a result of the transition to Normal will be executed sequentially. Thus, *OPC? or *WAI following CAL:AUTO ON will return when the alignment processing is complete.

The presence of an external signal may interfere with the RF portion of the alignment. If so, the Error Condition “Align skipped: 50 MHz interference” or

“Align skipped: 4.8 GHz interference” is reported, and bit 11 is set in the

Status Questionable Calibration register. After the interfering signal is removed, subsequent alignment of the RF will clear the condition, and clear bit 11 in the Status Questionable Calibration register.

Normal

An interfering user signal may prevent automatic alignment of the RF subsystem. If this occurs, the Error Condition “Align skipped: 50 MHz interference” or “Align skipped: 4.8 GHz interference” is reported, the Status

Questionable Calibration bit 11 is set, and the alignment proceeds. When a subsequent alignment of the RF subsystem succeeds, either by the next cycle of automatic alignment or from an

Align Now

,

RF

, the Error Condition and

Status Questionable Calibration bit 11 are cleared.

Prior to A.02.00

Initial S/W Revision

Partial

Auto Align

,

Partial

disables the full automatic alignment and the maintenance of warranted operation for the benefit of improved measurement throughput. Accuracy is retained for the Resolution Bandwidth filters and the IF Passband which is critical to FFT accuracy, demodulation, and many measurement applications. With Auto Align set to

Partial

, you are now responsible for maintaining warranted operation by updating the alignments when they expire. The

Auto Align

,

Alert

mechanism will notify you when alignments have expired. One solution to expired alignments is to perform the

Align All

,

Now

operation. Another is to return the

Auto Align

selection to

Normal

.

Auto Align

,

Partial

is recommended for measurements where the throughput is so important that a few percent of improvement is more valued than an increase in the accuracy errors of a few tenths of a decibel. One good application of

Auto Align

,

Partial

would be an automated environment where the alignments can be called during overhead time when the device-under-test is exchanged.

When

Auto Align

,

Partial

is selected the elapsed time counter begins for Auto Align Off time.

194

System Functions

System

When

Auto Align

,

Partial

is selected the Settings Panel indicates ALIGN PARTIAL with a warning icon. The warning icon is to inform the operator that they are responsible for maintaining the warranted operation of the instrument

Key Path

Mode

Example

Notes

Readback Text

Initial S/W Revision

System, Alignments, Auto Align

All

:CAL:AUTO PART

Auto Align Partial begins the elapsed time counter for Auto Align Off time.

Partial

Prior to A.02.00

Off

Auto Align

,

Off

disables automatic alignment and the maintenance of warranted operation, for the benefit of maximum measurement throughput. With Auto Align set to

Off

, you are now responsible for maintaining warranted operation by updating the alignments when they expire. The

Auto Align

,

Alert

mechanism will notify you when alignments have expired. One solution to expired alignments is to perform the

Align All

,

Now

operation. Another is to return the

Auto Align

selection to

Normal

.

The

Auto Align

,

Off

setting is rarely the best choice, because

Partial

gives almost the same improvement in throughput while maintaining the warranted performance for a much longer time. The

Off

choice is intended for unusual circumstances such as the measurement of radar pulses where you might like the revisit time to be as consistent as possible.

When

Auto Align

,

Off

is selected the Auto Align Off time is initialized and the elapsed time counter begins.

When

Auto Align

,

Off

is selected the Settings Panel indicates ALIGN OFF with a warning icon. The warning icon is to inform the operator that they are responsible for maintaining the warranted operation of the instrument:

Key Path

Mode

Example

Notes

Couplings

Readback Text

Initial S/W Revision

System, Alignments, Auto Align

All

:CAL:AUTO OFF

Auto Align Off begins the elapsed time counter for Auto Align Off time.

Auto Align is set to Off if Restore Align Data is invoked.

Off

Prior to A.02.00

All but RF

Auto Align

,

All but RF

, configures automatic alignment to include or exclude the RF subsystem.

(Eliminating the automatic alignment of the RF subsystem prevents the input impedance from changing.

The normal input impedance of 50 ohms can change to an open circuit when alignments are being used.

Some devices under test do not behave acceptably under such circumstances, for example by showing

195

System Functions

System

instability.) When

Auto Align

,

All but RF ON

is selected, the operator is responsible for performing an

Align Now

,

RF

when RF-related alignments expire. The

Auto Align

,

Alert

mechanism will notify the operator to perform an

Align Now

,

All

when the combination of time and temperature variation is exceeded.

When

Auto Align

,

All but RF ON

is selected the Settings Panel indicates ALIGN AUTO/NO RF with a warning icon (warning icon is intended to inform the operator they are responsible for the maintaining the RF alignment of the instrument):

Key Path

Mode

Remote Command

Example

Preset

Preset

State Saved

Readback Text

Initial S/W Revision

System, Alignments, Auto Align

All

:CALibration:AUTO:MODE ALL|NRF

:CALibration:AUTO:MODE?

:CAL:AUTO:MODE NRF

ALL

This is unaffected by Preset but is set to ALL on a “Restore System

Defaults->Align”.

No

RF or NRF

Prior to A.02.00

Alert

The instrument will signal an Alert when conditions exist such that you will need to perform a full alignment (for example,

Align Now

,

All

). The Alert can be configured in one of four settings;

Time &

Temperature

,

24 hours

,

7 days

, or

None

. A confirmation is required when a selection other than

Time

& Temperature

is chosen. This prevents accidental deactivation of alerts.

With

Auto Align

set to

Normal

, the configuration of

Alert

is not relevant because the instrument’s software maintains the instrument in warranted operation.

Key Path

Mode

Remote Command

Example

Notes

Preset

Preset

System, Alignments, Auto Align

All

:CALibration:AUTO:ALERt TTEMperature|DAY|WEEK|NONE

:CALibration:AUTO:ALERt?

:CAL:AUTO:ALER TTEM

The alert that alignment is needed is the setting of bit 14 in the Status

Questionable Calibration register.

TTEMperature

This is unaffected by Preset but is set to TTEMperature on a “Restore System

Defaults->Align”.

196

System Functions

System

State Saved

Status Bits/OPC dependencies

No

The alert is the Error Condition “Align Now, All required” and bit 14 is set in the Status Questionable Calibration register.

Prior to A.02.00

Initial S/W Revision

Time & Temperature

With Auto Align Alert set to

Time & Temperature

the instrument will signal an alert when alignments expire due to the combination of the passage of time and changes in temperature. The alert is the Error Condition “Align Now,

All required”. If this choice for Alert is selected, the absence of an alert means that the analyzer alignment is sufficiently up-to-date to maintain warranted accuracy.

Key Path

Mode

Example

Readback Text

Status Bits/OPC dependencies

Initial S/W Revision

System, Alignments, Auto Align, Alert

All

:CAL:AUTO:ALER TTEM

Time & Temp

Bit 14 is set in the Status Questionable Calibration register.

Prior to A.02.00

24 hours

With Auto Align Alert set to

24 Hours

the instrument will signal an alert after a time span of 24 hours since the last successful full alignment (for example,

Align Now

,

All

or completion of a full Auto Align). You may choose this selection in an environment where the temperature is stable on a daily basis at a small risk of accuracy errors in excess of the warranted specifications. The alert is the Error Condition “Align Now, All required”.

For front-panel operation , confirmation is required to transition into this setting of Alert. The confirmation dialog is:

No confirmation is required when Alert is configured through a remote command.

Key Path

Mode

Example

System, Alignments, Auto Align, Alert

All

:CAL:AUTO:ALER DAY

197

System Functions

System

Readback Text

Status Bits/OPC dependencies

Initial S/W Revision

24 hours

Bit 14 is set in the Status Questionable Calibration register.

Prior to A.02.00

7 days

With Auto Align Alert is set to

7 days

the instrument will signal an alert after a time span of 168 hours since the last successful full alignment (for example,

Align Now

,

All

or completion of a full Auto Align). You may choose this selection in an environment where the temperature is stable on a weekly basis, at a modest risk of accuracy degradations in excess of warranted performance. The alert is the Error Condition “Align Now, All required”.

For front panel operation, confirmation is required for the customer to transition into this setting of Alert. The confirmation dialog is:

No confirmation is required when Alert is configured through a remote command.

Key Path

Mode

Example

Readback Text

Status Bits/OPC dependencies

Initial S/W Revision

System, Alignments, Auto Align, Alert

All

:CAL:AUTO:ALER WEEK

7 days

Bit 14 is set in the Status Questionable Calibration register.

Prior to A.02.00

None

With Auto Align Alert set to

None

the instrument will not signal an alert. This is provided for rare occasions where you are making a long measurement which cannot tolerate Auto Align interruptions, and must have the ability to capture a screen image at the end of the measurement without an alert posted to the display. Agilent does not recommends using this selection in any other circumstances, because of the risk of accuracy performance drifting well beyond expected levels without the operator being informed.

For front panel operation, confirmation is required to transition into this setting of Alert. The confirmation dialog is:

198

System Functions

System

No confirmation is required when Alert is configured through a remote command.

Key Path

Mode

Example

Initial S/W Revision

System, Alignments, Auto Align, Alert

All

:CAL:AUTO:ALER NONE

Prior to A.02.00

Align Now

Accesses alignment processes that are immediate action operations. They perform complete operations and run until they are complete.

Key Path

Initial S/W Revision

System, Alignments

Prior to A.02.00

All

Immediately executes an alignment of all subsystems. The instrument stops any measurement currently underway, performs the alignment, then restarts the measurement from the beginning (similar to pressing the

Restart

key).

If an interfering user signal is present at the RF Input, the alignment is performed on all subsystems except the RF. After completion, the Error Condition “Align skipped: 50 MHz interference” or “Align skipped: 4.8 GHz interference” is set. In addition the Error Condition “Align Now, RF required” is set, and bits 11 and 12 are set in the Status Questionable Calibration register.

The query form of the remote commands (:CALibration[:ALL]? or *CAL?) invokes the alignment of all subsystems and returns a success or failure value. An interfering user signal is not grounds for failure; if the alignment was able to succeed on all portions but unable to align the RF because of an interfering signal, the resultant will be the success value.

Successful completion of

Align Now

,

All

will clear the “Align Now, All required” Error Condition, and clear bit 14 in the Status Questionable Calibration register. It will also begin the elapsed time counter for

Last Align Now, All Time, and capture the Last Align Now, All Temperature.

If the Align RF subsystem succeeded in aligning (no interfering signal present), the elapsed time counter begins for Last Align Now, RF Time, and the temperature is captured for the Last Align Now, RF

199

System Functions

System

Temperature. In addition the Error Conditions “Align skipped: 50 MHz interference” and “Align skipped: 4.8 GHz interference” are cleared, the Error Condition “Align Now, RF required” is cleared, and bits 11 and 12 are cleared in the Status Questionable Calibration register

Align Now

,

All

can be interrupted by pressing the Cancel (ESC) front-panel key or remotely with Device

Clear followed by the :ABORt SCPI command. When this occurs the Error Condition “Align Now, All required” is set, and bit 14 is set in the Status Questionable Condition register. This is because new alignment data may be employed for an individual subsystem, but not a cohesive set of data for all subsystems.

In many cases, you might find it more convenient to change alignments to

Normal

, instead of executing

Align Now

,

All

. When the Auto Align process transitions to

Normal

, the analyzer will immediately start to update only the alignments that have expired, thus efficiently restoring the alignment process.

Key Path

Mode

Remote Command

Example

Notes

Couplings

Status Bits/OPC dependencies

System, Alignments, Align Now

All

:CALibration[:ALL]

:CALibration[:ALL]?

:CAL

:CALibration[:ALL]? returns 0 if successful

:CALibration[:ALL]? returns 1 if failed

:CALibration[:ALL]? is the same as *CAL?

While Align Now, All is performing the alignment, bit 0 in the Status

Operation register is set. Completion, or termination, will clear bit 0 in the

Status Operation register.

This command is sequential; it must complete before further SCPI commands are processed. Interrupting the alignment from remote is accomplished by invoking Device Clear followed by the :ABORt command.

Successful completion will clear bit 14 in the Status Questionable Calibration register.

An interfering user signal is not grounds for failure of Align Now, All.

However, bits 11 and 12 are set in the Status Questionable Calibration register to indicate Align Now, RF is required.

An interfering user supplied signal will result in the instrument requiring an

Align Now, RF with the interfering signal removed.

Initializes the time for the Last Align Now, All Time.

Records the temperature for the Last Align Now, All Temperature.

If Align RF component succeeded, initializes the time for the Last Align Now,

RF Time.

If Align RF component succeeded, records the temperature for the Last Align

Now, RF Temperature.

Bits 11, 12, or 14 may be set in the Status Questionable Calibration register.

200

System Functions

System

Initial S/W Revision Prior to A.02.00

Mode

Remote Command

Example

Notes

All

*CAL?

*CAL?

*CAL? returns 0 if successful

*CAL? returns 1 if failed

:CALibration[:ALL]? is the same as *CAL?

See additional remarks described with :CALibration[:ALL]?

Everything about :CALibration[:ALL]? is synonymous with *CAL? including all conditions, status register bits, and couplings

Prior to A.02.00

Initial S/W Revision

All but RF

Immediately executes an alignment of all subsystems except the RF subsystem. The instrument will stop any measurement currently underway, perform the alignment, and then restart the measurement from the beginning (similar to pressing the

Restart

key). This can be used to align portions of the instrument that are not impacted by an interfering user input signal.

This operation might be chosen instead of

All

if you do not want the device under test to experience a large change in input impedance, such as a temporary open circuit at the analyzer input.

The query form of the remote commands (:CALibration:NRF?) will invoke the alignment and return a success or failure value.

Successful completion of

Align Now

,

All but RF

will clear the “Align Now, All required” Error

Condition, and clear bit 14 in the Status Questionable Calibration register. If “Align Now, All required” was in effect prior to executing the All but RF, the Error Condition “Align Now, RF required” is asserted and bit 12 in the Status Questionable Calibration register is set. It will also begin the elapsed time counter for Last Align Now, All Time, and capture the Last Align Now, All Temperature.

Align Now

,

All but RF

can be interrupted by pressing the Cancel (ESC) front-panel key or remotely with

Device Clear followed by the :ABORt SCPI command. When this occurs the Error Condition “Align

Now, All required” is set, and bit 14 is set in the Status Questionable Condition register. This is because new alignment data may be used for an individual subsystem, but not a full new set of data for all subsystems.

Key Path

Mode

Remote Command

Example

System, Alignments, Align Now

All

:CALibration:NRF

:CALibration:NRF?

:CAL:NRF

201

System Functions

System

Notes

Couplings

:CALibration:NRF? returns 0 if successful

:CALibration:NRF? returns 1 if failed

While Align Now, All but RF is performing the alignment, bit 0 in the Status

Operation register is set. Completion, or termination, will clear bit 0 in the

Status Operation register.

This command is sequential; it must complete before further SCPI commands are processed. Interrupting the alignment from remote is accomplished by invoking Device Clear followed by the :ABORt command.

Successful completion will clear bit 14 in the Status Questionable Calibration register and set bit 12 if invoked with “Align Now, All required”.

Initializes the time for the Last Align Now, All Time.

Records the temperature for the Last Align Now, All Temperature.

Bits 12 or 14 may be set in the Status Questionable Calibration register.

Prior to A.02.00

Status Bits/OPC dependencies

Initial S/W Revision

RF

Immediately executes an alignment of the RF subsystem. The instrument stops any measurement currently underway, performs the alignment, then restarts the measurement from the beginning (similar to pressing the

Restart

key).

This operation might be desirable if the alignments had been set to not include RF alignments, or if previous RF alignments could not complete because of interference which has since been removed.

If an interfering user signal is present at the RF Input, the alignment will terminate and raise the Error

Condition “Align skipped: 50 MHz interference” or “Align skipped: 4.8 GHz interference”, and Error

Condition “Align Now, RF required”. In addition, bits 11 and 12 will be set in the Status Questionable

Calibration register.

The query form of the remote commands (:CALibration:RF?) will invoke the alignment of the RF subsystem and return a success or failure value. An interfering user signal is grounds for failure.

A failure encountered during alignment will set the Error Condition “Align RF failed” and set bit 3 in the

Status Questionable Calibration register.

Successful completion of

Align Now

,

RF

clears the Error Conditions “Align skipped: 50 MHz interference” and “Align skipped: 4800 MHz interference” and the Error Conditions “Align RF failed” and “Align Now, RF required”, and clears bits 3, 11, and 12 in the Status Questionable Calibration register. It will also begin the elapsed time counter for Last Align Now, RF Time, and capture the Last

Align Now, RF Temperature.

Align Now

,

RF

can be interrupted by pressing the Cancel (ESC) front-panel key or remotely with Device

Clear followed by the :ABORt SCPI command. When this occurs, the Error Condition “Align Now, RF required” is set, and bit 12 is set in the Status Questionable Condition register. None of the new alignment data is used.

Key Path

Mode

System, Alignments, Align Now

All

202

System Functions

System

Remote Command

Example

Notes

Couplings

:CALibration:RF

:CALibration:RF?

:CAL:RF

:CALibration:RF? returns 0 if successful

:CALibration:RF? returns 1 if failed (including interfering user signal)

While Align Now, RF is performing the alignment, bit 0 in the Status

Operation register is set. Completion, or termination, will clear bit 0 in the

Status Operation register.

This command is sequential; it must complete before further SCPI commands are processed. Interrupting the alignment from remote is accomplished by invoking Device Clear followed by the :ABORt command.

Successful completion will clear bits 3, 11, and 12 in the Status Questionable

Calibration register.

A failure encountered during alignment will set the Error Condition “Align

RF failed” and set bit 3 in the Status Questionable Calibration register.

An interfering user signal will result in bits 11 and 12 to be set in the Status

Questionable Calibration register to indicate Align Now, RF is required.

An interfering user supplied signal will result in the instrument requiring an

Align Now, RF with the interfering signal removed.

Initializes the time for the Last Align Now, RF Time.

Records the temperature for the Last Align Now, RF Temperature.

Bits 11, 12, or 14 may be set in the Status Questionable Calibration register.

Prior to A.02.00

Status Bits/OPC dependencies

Initial S/W Revision

Show Alignment Statistics

Shows alignment information you can use to ensure that the instrument is operating in a specific manner.

The Show Alignment Statistics screen is where you can view time and temperature information.

Values which are displayed are only updated when the Show Alignment Statistics screen is invoked, they are not updated while the Show Alignment Statistics screen is being displayed. The remote commands which access this information obtain current values.

An example of the Show Alignment Statistics screen would be similar to:

203

System Functions

System

A successful Align Now, RF will set the Last Align RF temperature to the current temperature, and reset the Last Align RF time. A successful Align Now, All or Align Now, All but RF will set the Last Align

Now All temperature to the current temperature, and reset the Last Align Now All time. A successful

Align Now, All will also reset the Last Align RF items if the RF portion of the Align Now succeeded.

Key Path

Mode

Notes

Initial S/W Revision

System, Alignments

All

The values displayed on the screen are only updated upon entry to the screen and not updated while the screen is being displayed.

Prior to A.02.00

Key Path

Mode

Remote Command

Example

Notes

State Saved

Initial S/W Revision

Key Path

Mode

Remote Command

Visual annotation in the Show Alignment Statistics screen

All

:SYSTem:PON:TIME?

:SYST:PON:TIME?

Value is the time since the most recent start-up in seconds.

No

Prior to A.02.00

Visual annotation in the Show Alignment Statistics screen

All

:CALibration:TEMPerature:CURRent?

204

Example

Notes

State Saved

Initial S/W Revision

Key Path

Mode

Remote Command

Example

Notes

State Saved

Initial S/W Revision

Key Path

Mode

Remote Command

Example

Notes

State Saved

Initial S/W Revision

Key Path

Mode

Remote Command

Example

Notes

State Saved

Initial S/W Revision

Key Path

System Functions

System

:CAL:TEMP:CURR?

Value is in degrees Centigrade.

Value is invalid if using default alignment data (Align Now, All required)

No

Prior to A.02.00

Visual annotation in the Show Alignment Statistics screen

All

:CALibration:TIME:LALL?

:CAL:TIME:LALL?

Value is the elapsed time, in seconds, since the last successful Align Now, All or Align Now, All but RF was executed.

No

Prior to A.02.00

Visual annotation in the Show Alignment Statistics screen

All

:CALibration:TEMPerature:LALL?

:CAL:TEMP:LALL?

Value is in degrees Centigrade at which the last successful Align Now, All or

Align Now, All but RF was executed.

No

Prior to A.02.00

Visual annotation in the Show Alignment Statistics screen

All

:CALibration:TIME:LRF?

:CAL:TIME:LRF?

Value is the elapsed time, in seconds, since the last successful Align Now, RF was executed, either individually or as a component of Align Now, All.

No

Prior to A.02.00

Visual annotation in the Show Alignment Statistics screen

205

System Functions

System

Mode

Remote Command

Example

Notes

State Saved

Initial S/W Revision

Key Path

Mode

Remote Command

Example

Notes

Dependencies

State Saved

Initial S/W Revision

Key Path

Mode

Remote Command

Example

Notes

Dependencies

State Saved

Initial S/W Revision

Key Path

Mode

Remote Command

Example

206

All

:CALibration:TEMPerature:LRF?

:CAL:TEMP:LRF?

Value is in degrees Centigrade at which the last successful Align Now, RF was executed, either individually or as a component of Align Now, All.

No

Prior to A.02.00

Visual annotation in the Show Alignment Statistics screen

All

:CALibration:TIME:LPReselector?

:CAL:TIME:LPR?

Value is date and time the last successful Characterize Preselector was executed. The date is separated from the time by a space character. Returns “” if no Characterize Preselector has ever been performed on the instrument.

In models that do not include preselectors, this command is not enabled and any attempt to set or query will yield an error.

No

Prior to A.02.00

Visual annotation in the Show Alignment Statistics screen

All

:CALibration:TEMPerature:LPReselector?

:CAL:TEMP:LPR?

Value is in degrees Centigrade at which the last successful Characterize

Preselector was executed.

In models that do not include preselectors, this command is not enabled and any attempt to set or query will yield an error.

No

Prior to A.02.00

Visual annotation in the Show Alignment Statistics screen

All

:CALibration:AUTO:TIME:OFF?

:CAL:AUTO:TIME:OFF?

System Functions

System

Notes

State Saved

Initial S/W Revision

Value is the elapsed time, in seconds, since Auto Align has been set to Off or

Off with Alert. The value is 0 if Auto Align is ALL or NORF.

No

Prior to A.02.00

Restore Align Defaults

Initializes the alignment user interface settings, not alignment data, to the factory default values. Align

Now, All must be executed if the value of the Timebase DAC results in a change.

For front panel operation, you are prompted to confirm action before setting the alignment parameters to factory defaults:

The parameters affected are:

Parameter

Timebase DAC

Timebase DAC setting

Auto Align State

Auto Align All but RF

Auto Align Alert

Key Path

Mode

Example

Notes

Initial S/W Revision

Setting

Calibrated

Calibrated value

Normal (if the instrument is not operating with default alignment data, Off otherwise)

Off

Time & Temperature

System, Alignments

All

:SYST:DEF ALIG

Alignment processing that results as the transition to Auto Alignment Normal will be executed sequentially; thus *OPC? or *WAI will wait until the alignment processing is complete.

Prior to A.02.00

207

System Functions

System

Backup and Restore Alignment Data

Alignment data for the instrument resides on the hard drive in a database. Agilent uses high quality hard drives; however it is highly recommended the alignment data be backed-up to storage outside of the instrument. Additionally, for customers who use multiple CPU Assemblies or multiple disk drives, the alignment that pertains to the instrument must be transferred to the resident hard drive after a CPU or hard drive is replaced. This utility facilitates backing-up and restoring the alignment data.

NOTE

This utility allows the operator to navigate to any location of the Windows file system. It is intended that the operator use a USB memory device or Mapped

Network Drive to backup the alignment data to storage outside of the instrument.

Backup or Restore Align Data…

Opens the utility for backing-up or restoring the alignment data.

Key Path

Initial S/W Revision

System, Alignments

A.02.00

Key Path

Mode

Remote Command

Example

Couplings

System, Alignments

All

:CALibration:DATA:DEFault

:CAL:DATA:DEF

Sets Auto Align to Off. Sets bit 14 in the Status Questionable Calibration register. The Error Condition “Align Now, All required” is set.

Prior to A.02.00

Initial S/W Revision

Alignment Data Wizard

The Backup or Restore Alignment Data wizard will guide you through the operation of backing-up or restoring the alignment data.

The following dialogue boxes operate without a mouse or external keyboard when you use the default file names.

208

System Functions

System

The backup screen will indicate the approximate amount of space required to contain the backup file.

The default file name will be AlignDataBackup_<model number>_<serial number>_<date in

YYYYMMDDHHMMSS>.bak.

For the N9030A the default backup location will be the internal F: drive which is a solid-state memory device

209

System Functions

System

located internally on the instrument.

Changing the drive letter will also modify the path displayed in the box below. When this step is first loaded, the drive drop-down is populated with connected drives which provide the user with write access. If there are many unreachable network drives connected to the instrument, this step can take a few seconds. If a USB drive is present, it will be selected by default. The path defaults to the AlignmentBackups folder, and a filename will be automatically created in the form of AlignDataBackup_<model>_<serial number>_<date><time>. When the

"Next >" button is pressed, the user will be prompted to create a new folder if the chosen path does not yet exist.

210

System Functions

System

The restore operation will check the validity of the restore file using the database's built-in file validation. If the restore file is corrupt, the existing alignment data will remain in use.

For the N9030A, the default restore location will be the internal F: drive which is a solid-state memory device

211

System Functions

System

located internally on the instrument. The default restore file will be the most recent file that matches the default backup file name format: AlignDataBackup_N9030A_<serial number>_<date>.bak

Changing the drive letter will also modify the path displayed in the box below. When this step is first loaded, the drive drop-down is populated with connected drives which provide the user with read access. The path defaults to the AlignBackups folder. The most recent *.bak file in the folder will also be selected by default.

212

System Functions

System

Perform Backup (Remote Command Only)

Invokes an alignment data backup operation to the provided Folder.

213

System Functions

System

NOTE

It is recommended that the Folder provided is outside of the instrument (USB or Mapped

Network Drive).

Remote Command:

Example:

:CALibration:DATA:BACKup <filename>

:CAL:DATA:BACK

"F:\AlignDataBackup_N9020A_US00000001_2008140100.bak"

A.02.00

Initial S/W Revision:

Perform Restore (Remote Command Only)

Invokes an alignment data restore operation from the provided filename.

Remote Command:

Example:

Initial S/W Revision:

:CALibration:DATA:RESTore <filename>

:CAL:DATA:REST "F:\

AlignDataBackup_N9020A_US00000001_2008140100.bak "

A.02.00

Advanced

Accesses alignment processes that are immediate action operations that perform operations that run until complete. Advanced alignments are performed on an irregular basis, or require additional operator interaction

Key Path

Initial S/W Revision

System, Alignments

Prior to A.02.00

Characterize Preselector (Only with Option 507, 508, 513, or 526)

The Preselector tuning curve drifts over temperature and time. Recognize that the

Amplitude

,

Presel

Center

function adjusts the preselector for accurate amplitude measurements at an individual frequency.

Characterize Preselector

improves the amplitude accuracy by ensuring the Preselector is approximately centered at all frequencies without the use of the

Amplitude

,

Presel Center

function.

Characterize Preselector

can be useful in situations where absolute amplitude accuracy is not of utmost importance, and the throughput savings or convenience of not performing a

Presel Center

is desired.

Presel Center

is required prior to any measurement for best (and warranted) amplitude accuracy.

Agilent recommends that the

Characterize Preselector

operation be performed yearly as part of any calibration, but performing this operation every three months can be worthwhile.

Characterize Preselector

immediately executes a characterization of the Preselector, which is a

YIG-tuned filter (YTF). The instrument stops any measurement currently underway, performs the characterization, then restarts the measurement from the beginning (similar to pressing the

Restart

key).

The query form of the remote commands (:CALibration:YTF?) will invoke the alignment of the YTF

214

System Functions

System

subsystem and return a success or failure value.

A failure encountered during alignment will set the Error Condition “Characterize YTF failed” and set bit 9 in the Status Questionable Calibration register.

Successful completion of

Advanced

,

Characterize Preselector

will clear the Error Condition

“Characterize YTF failed”, and clear bit 9 in the Status Questionable Calibration register. It will also begin the elapsed time counter for Last Characterize Preselector Time, and capture the Last Characterize

Preselector Temperature.

The last Characterize Preselector Time and Temperature must survive across the power cycle as this operation is performed infrequently.

Advanced

,

Characterize Preselector

can be interrupted by pressing the Cancel (ESC) front-panel key or remotely with Device Clear followed by the :ABORt SCPI command. None of the new characterization data is then used.

Remote Command

Example

Initial S/W Revision

Key Path

Mode

Notes

Dependencies

Couplings

:CALibration:YTF

:CALibration:YTF?

:CAL:YTF

Prior to A.02.00

System, Alignments, Advanced

All

:CALibration:YTF? returns 0 if successful

:CALibration:YTF? returns 1 if failed (including interfering user signal)

While Advanced, Characterize Preselector is performing the alignment, bit 0 in the Status Operation register is set. Completion, or termination, will clear bit 0 in the Status Operation register.

This command is sequential; it must complete before further SCPI commands are processed. Interrupting the alignment from remote is accomplished by invoking Device Clear followed by the :ABORt command.

Successful completion will clear bit 9 in the Status Questionable Calibration register.

A failure encountered during alignment will set the Error Condition

“Characterize Preselector failed” and set bit 9 in the Status Questionable

Calibration register.

For Option 507, 508, 513, and 526 only.

This key does not appear in models that do not contain preselectors. In these models the SCPI command is accepted without error but no action is taken.

Initializes the time for the Last Characterize Preselector Time.

Records the temperature for the Last Characterize Preselector Temperature.

Timebase DAC

Allows control of the internal 10 MHz reference oscillator timebase. This may be used to adjust for

215

System Functions

System

minor frequency alignment between the signal and the internal frequency reference. This adjustment has no effect if the instrument is operating with an External Frequency Reference.

If the value of the Timebase DAC changes (by switching to Calibrated from User with User set to a different value, or in User with a new value entered) an alignment may be necessary. The alignment system will take appropriate action; which will either invoke an alignment or cause an Alert.

Key Path

Mode

Remote Command

Example

Notes

Preset

Preset

State Saved

Initial S/W Revision

System, Alignments

All

:CALibration:FREQuency:REFerence:MODE CALibrated|USER

:CALibration:FREQuency:REFerence:MODE?

:CAL:FREQ:REF:MODE CAL

If the value of the timebase is changed the alignment system automatically performs an alignment or alerts that an alignment is due.

If the value of the timebase is changed the alignment system automatically performs an alignment or alerts that an alignment is due.

CAL

This is unaffected by Preset but is set to CALibrated on a “Restore System

Defaults->Align”.

No

Prior to A.02.00

Calibrated

Sets the Timebase DAC to the value established during factory or field calibration. The value displayed on the menu key is the calibrated value.

Key Path

Mode

Example

Readback Text

Initial S/W Revision

System, Alignments, Timebase DAC

All

:CAL:FREQ:REF:MODE CAL

[xxx] < where xxx is the calibrated value

Prior to A.02.00

User

Allows setting the Timebase DAC to a value other than the value established during the factory or field calibration. The value displayed on the menu key is the calibrated value.

Key Path

Mode

Example

System, Alignments, Timebase DAC

All

:CAL:FREQ:REF:MODE USER

216

System Functions

System

Readback Text

Initial S/W Revision

Key Path

Mode

Remote Command

Example

Notes

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision xxx < where xxx is the Timebase DAC setting

Prior to A.02.00

System, Alignments, Timebase DAC

All

:CALibration:FREQuency:REFerence:FINE <integer>

:CALibration:FREQuency:REFerence:FINE?

:CAL:FREQ:REF:FINE 8191

If the value of the timebase is changed the alignment system automatically performs an alignment or alerts that an alignment is due.

Setting :CAL:FREQ:REF:FINE sets :CAL:FREQ:REF:MODE USER

This is unaffected by Preset but is set to the factory setting on a “Restore

System Defaults->Align”.

No

0

16383

Prior to A.02.00

Remote Command:

Example:

Notes:

Couplings:

Initial S/W Revision:

:CALibration:FREQuency:REFerence:COARse <integer>

:CALibration:FREQuency:REFerence:COARse?

:CAL:FREQ:REF:COAR 8191

This is an alias for CAL:FREQ:REF:FINE any change to COARse is reflected in FINE and vice-versa. See CAL:FREQ:REF:FINE for description of functionality.

Setting :CAL:FREQ:REF:COAR sets :CAL:FREQ:REF:MODE USER

Prior to A.02.00

I/O Config

Activates a menu for identifying and changing the I/O configuration for remote control.

Key Path

Initial S/W Revision

System

Prior to A.02.00

217

System Functions

System

GPIB

Activates a menu for configuring the GPIB I/O port.

Key Path

Initial S/W Revision

GPIB Address

Select the GPIB remote address.

System, I/O Config

A.02.00

Key Path

Mode

Remote Command

Example

Notes

Preset

Preset

State Saved

Range

Initial S/W Revision

System, I/O Config, GPIB

All

:SYSTem:COMMunicate:GPIB[1][:SELF]:ADDRess <integer>

:SYSTem:COMMunicate:GPIB[1][:SELF]:ADDRess?

:SYST:COMM:GPIB:ADDR 17

Changing the Address on the GPIB port requires all further communication to use the new address.

18

This is unaffected by Preset but is set to 18 on a “Restore System

Defaults->Misc”

No

0 to 30

Prior to A.02.00

GPIB Controller

Sets the GPIB port into controller or device mode. In the normal state, GPIB controller is disabled, which allows the analyzer to be controlled by a remote computer. When GPIB Controller is enabled, the instrument can run software applications that use the instrument's computer as a GPIB controller; controlling devices connected to the instrument's GPIB port.

NOTE

When GPIB Controller is enabled, the analyzer application itself cannot be controlled over GPIB. In this case it can easily be controlled via LAN or USB.

The GPIB port cannot be a controller and device at the same time. Only one controller can be active on the GPIB bus at any given time. If the analyzer is the controller, an external PC cannot be a controller.

To control the instrument from the software that is performing GPIB controller operation, you can use an internal TCP/IP connection to the analyzer application. Use the address

TCPIP0::localhost::inst0::INSTR to send SCPI commands to the analyzer application.

Key Path

System, I/O Config, GPIB

218

System Functions

System

Mode

Scope

Remote Command

Example

Notes

Preset

Preset

State Saved

Range

Initial S/W Revision

All

Mode Global

:SYSTem:COMMunicate:GPIB[1][:SELF]:CONTroller[:ENABle]

ON|OFF|0|1

:SYSTem:COMMunicate:GPIB[1][:SELF]:CONTroller[:ENABle]?

:SYST:COMM:GPIB:CONT ON ! Will set GPIB port to Controller

When the instrument becomes the Controller bit 0 in the Standard Event

Status Register is set (and when the instrument relinquishes Controller capability bit 0 is cleared in the Standard Event Status Register).

OFF

This is unaffected by Preset but is set to OFF on a “Restore System

Defaults->Misc”

No

Disabled|Enabled

A.02.00

Disabled

Disables the GPIB Controller capability, this is the default (or normal) setting.

Key Path

Example

Initial S/W Revision

Enabled

Enables the GPIB Controller capability.

System, I/O Config, GPIB, GPIB Controller

:SYST:COMM:GPIB:CONT OFF ! Will set GPIB port to Device

A.02.00

Key Path

Example

Initial S/W Revision

System, I/O Config, GPIB, GPIB Controller

:SYST:COMM:GPIB:CONT ON ! Will set GPIB port to Controller

A.02.00

SCPI LAN

Activates a menu for identifying and changing the SCPI over a LAN configuration. There are a number of different ways to send SCPI remote commands to the instrument over LAN. It can be a problem to have multiple users simultaneously accessing the instrument over the LAN. These keys limit that somewhat by disabling the telnet, socket, and/or SICL capability.

Key Path

Initial S/W Revision

System, I/O Config

Prior to A.02.00

219

System Functions

System

SCPI Telnet

Turns the SCPI LAN telnet capability On or Off allowing you to limit SCPI access over LAN through telnet.

Key Path

Mode

Remote Command

Example

Preset

Preset

State Saved

Range

Initial S/W Revision

System, I/O Config, SCPI LAN

All

:SYSTem:COMMunicate:LAN:SCPI:TELNet:ENABle OFF|ON|0|1

:SYSTem:COMMunicate:LAN:SCPI:TELNet:ENABle?

:SYST:COMM:LAN:SCPI:TELN:ENAB OFF

ON

This is unaffected by Preset but is set to ON with a “Restore System

Defaults->Misc”

No

On | Off

Prior to A.02.00

SCPI Socket

Turns the capability of establishing Socket LAN sessions On or Off. This allows you to limit SCPI access over LAN through socket sessions.

Key Path

Mode

Remote Command

Example

Preset

Preset

State Saved

Range

Initial S/W Revision

System, I/O Config, SCPI LAN

All

:SYSTem:COMMunicate:LAN:SCPI:SOCKet:ENABle OFF|ON|0|1

:SYSTem:COMMunicate:LAN:SCPI:SOCKet:ENABle?

:SYST:COMM:LAN:SCPI:SOCK:ENAB OFF

ON

This is unaffected by a Preset but is set to ON with a “Restore System

Defaults->Misc”

No

On | Off

Prior to A.02.00

SCPI Socket Control Port (Remote Command Only)

Returns the TCP/IP port number of the control socket associated with the SCPI socket session. This query enables you to obtain the unique port number to open when a device clear is to be sent to the instrument. Every time a connection is made to the SCPI socket, the instrument creates a peer control socket. The port number for this socket is random. The user must use this command to obtain the port number of the control socket. To force a device clear on this socket, open the port and send the string

220

System Functions

System

“DCL ” to the instrument.

If this SCPI command is sent to a non SCPI Socket interface, then 0 is returned.

Mode

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

All

:SYSTem:COMMunicate:LAN:SCPI:SOCKet:CONTrol?

:SYST:COMM:LAN:SCPI:SOCK:CONT?

This is unaffected by Preset or “Restore System Defaults->Misc”.

No

0 to 65534

Prior to A.02.00

SICL Server

Turns the SICL server capability On or Off, enabling you to limit SCPI access over LAN through the

SICL server. (SICL IEEE 488.2 protocol.)

Parameter

Maximum Connections

Instrument Name

Instrument Logical Unit

Description

The maximum number of connections that can be accessed simultaneously

The name (same as the remote SICL address) of your analyzer

The unique integer assigned to your analyzer when using

SICL LAN

Emulated GPIB Name

Emulated GPIB Logical Unit The unique integer assigned to your device when it is being controlled using SICL LAN

Emulated GPIB Address

The name (same as the remote SICL address) of the device used when communicating with your analyzer

The emulated GPIB address assigned to your transmitter tester when it is a SICL server (the same as your GPIB address)

Setting

5 inst0

8 gpib7

8

18

Key Path

Mode

Remote Command

Example

Preset

Preset

System, I/O Config, SCPI LAN

All

:SYSTem:COMMunicate:LAN:SCPI:SICL:ENABle OFF|ON|0|1

:SYSTem:COMMunicate:LAN:SCPI:SICL:ENABle?

:SYST:COMM:LAN:SCPI:SICL:ENAB OFF

ON

This is unaffected by Preset, but is set to ON with a “Restore System

Defaults->Misc”

221

System Functions

System

State Saved

Range

Initial S/W Revision

No

On | Off

Prior to A.02.00

Reset Web Password

The embedded web server contains certain capability which are password protected; modifying the LAN configuration of the instrument, and access to web pages that can change the settings of the instrument.

The default password from the factory is ‘agilent’ (without the quotes). The control provided here is the means to set the web password as the user desires, or to reset the password to the factory default.

Selecting Reset web password brings up a control for resetting the password as the user desires, or to the factory default. A keyboard is required to change the password from the factory default of ‘agilent’ or to set a new password that contains alphabetic characters. The control is:

If this control is entered without an external keyboard or mouse connected, you can cancel the control by pressing the Cancel (ESC) front-panel key.

Key Path

Mode

Initial S/W Revision

System, I/O Config

All

Prior to A.02.00

Query USB Connection (Remote Command Only)

Enables you to determine the speed of the USB connection.

Mode

Remote Command

Example

All

:SYSTem:COMMunicate:USB:CONNection?

:SYST:COMM:USB:CONN?

222

System Functions

System

Notes

State Saved

Range

Initial S/W Revision

NONE – Indicates no USB connection has been made.

LSPeed – Indicates a USB low speed connection (1.5 Mbps). Note: this is reserved for future use, the T+M488 protocol is not supported on low speed connections.

HSPeed – Indicates that a USB high speed connection (480 Mbps) has been negotiated.

FSPeed – Indicates that a USB full speed connection (12 Mbps) has been negotiated.

No

NONE|LSPeed|HSPeed|FSPeed

Prior to A.02.00

USB Connection Status (Remote Command Only)

Enables you to determine the current status of the USB connection.

Mode

Remote Command

Example

Notes

State Saved

Range

Initial S/W Revision

All

:SYSTem:COMMunicate:USB:STATus?

:SYST:COMM:USB:STAT?

SUSPended – Indicates that the USB bus is currently in its suspended state.

The bus is in the suspended state when:

The bus is not connected to any controller

The controller is currently powered off

The controller has explicitly placed the USB device into the suspended state.

When in the suspended state, no USB activity, including start of frame packets are received.

ACTive – Indicates that the USB device is in the active state. When the device is in the active state, it is receiving periodic start of frames but it isn’t necessarily receiving or transmitting data.

No

SUSPended|ACTive

Prior to A.02.00

USB Packet Count (Remote Command Only)

Enables you to determine the number of packets received and transmitted on the USB bus.

Mode

Remote Command

Example

All

:SYSTem:COMMunicate:USB:PACKets?

:SYST:COMM:USB:PACK?

223

System Functions

System

Notes

State Saved

Initial S/W Revision

Two integers are returned. The first is the number of packets received since application invocation, the second is the number of packets transmitted since application invocation. If no packets have been received or transmitted the response is 0,0.

The packet count is initialized to 0,0 when the instrument application is started.

No

Prior to A.02.00

LXI

Opens a menu that allows you to access the various LXI configuration properties.

Key Path

Initial S/W Revision

LAN Reset

Resets the LAN connection.

System, I/O Config

Prior to A.02.00

Key Path

Initial S/W Revision

System, I/O Config, LXI

Prior to A.02.00

LXI Domain

The instrument only receives LXI LAN Events sent by members of the same LXI Domain. Conversely,

LXI Output LAN Events sent by the instrument can only be received by members of the same LXI

Domain. This is not the same as the IEEE 1588 domain (see

“Domain (Remote Command Only)” on page 240 ).

Key Path

Remote Command

Example

Preset

Preset

State Saved

Range

Initial S/W Revision

System, I/O Config, LXI

:LXI:EVENt:DOMain <intDomain>

:LXI:EVENt:DOMain?

:LXI:EVEN:DOM 128

:LXI:EVEN:DOM?

0

Not affected by a Preset. The default value of "0" can be restored by pressing

Restore Defs, Input/Output Settings

Saved in instrument state.

0–255

Prior to A.02.00

224

System Functions

System

LXI Output LAN Events

The device can be configured to send LXI LAN Events as the instrument’s state changes. Specifically, it can notify other devices as the status signals WaitingForTrigger, Sweeping, Measuring,

OperationComplete, and Recalling transition. Additionally, Output LAN Events can be sent in response to the receipt of any of the Input LAN Events.

This is the entry point for the LXI Output LAN Event system. This key branches to a list of events that can be sent out on the LAN in response to instrument events.

Key Path

Initial S/W Revision

System, I/O Config, LXI

Prior to A.02.00

Disable All

This command causes the Enable property of all members of the LXI Output LAN Event List to be set to OFF.

Key Path

Remote Command

Example

Initial S/W Revision

System, I/O Config, LXI, LXI Output LAN Events

:LXI:EVENt[:OUTPut]:LAN:DISable:ALL

:LXI:EVEN:LAN:DIS:ALL

Prior to A.02.00

Output LAN Event List

This is the list of LXI Output LAN events that can be sent in response to an instrument event such as sweeping or waiting for a trigger. Each member of this list has a key in the LXI Output LAN Events panel. The list can grow and shrink in response to Add and Remove commands respectively. New pages must be added and removed automatically as the list size changes. Only the first 14 characters of an LXI Output LAN Event name are displayed on the key.

Key Path

Remote Command

Example

Preset

Preset

System, I/O Config, LXI, LXI Output LAN Events

:LXI:EVENt[:OUTPut]:LAN:LIST?

:LXI:EVEN:LAN:LIST?

Returns the complete list of Output LAN Events which is, at minimum:

“LAN0”, “LAN1”, “LAN2”, “LAN3”, “LAN4”, “LAN5”, “LAN6”, “LAN7”,

“WaitingForTrigger”, “Measuring”, “Sweeping”, OperationComplete”,

“Recalling”

“LAN0”, “LAN1”, “LAN2”, “LAN3”, “LAN4”, “LAN5”, “LAN6”, “LAN7”,

“WaitingForTrigger”, Measuring”, Sweeping”, OperationComplete”,

“Recalling”

Not affected by a Preset. The default values can be restored by pressing

Restore Defs, Input/Output Settings.

Preset/Default values: “LAN0”, “LAN1”, “LAN2”, “LAN3”, “LAN4”,

“LAN5”, “LAN6”, “LAN7”, “WaitingForTrigger”, Measuring”, Sweeping”,

OperationComplete”, “Recalling”

225

System Functions

System

State Saved

Readback Text

Saved in instrument state.

Displays the value of the LXI Output LAN Event Enabled parameter (Enabled

| Disabled). Also displays the value of the LXI Output LAN Event Source parameter (WaitingForTrig | Sweeping | Measuring | OpComplete | Recalling)

Prior to A.02.00

Initial S/W Revision

Add (Remote Command Only)

Adds the provided string to the list of possible LAN events to output as a response to instrument events. As new

LAN events are added, keys are generated in the LXI Output LAN Events menu. New key panels are generated as the number of possible LAN events increases past a multiple of six, and the “More” keys are updated to reflect the new number of key panels in the LXI Output LAN Events menu.

Remote Command:

Example:

Notes:

State Saved:

Range:

Initial S/W Revision:

:LXI:EVENt[:OUTPut]:LAN:ADD “LANEVENT”

:LXI:EVEN:LAN:ADD “LANEVENT”

The maximum length of the string is 16 characters.

Longer strings are concatenated and added to the LXI Output LAN Event list.

No event is added if the LAN Event already exists.

No

Uppercase, Lowercase, Numeric, Symbol except for comma or semicolon

Prior to A.02.00

Remove (Remote Command Only)

Removes the provided string from the list of possible LAN events to output as a response to instrument events. As new LAN events are removed, keys are removed from the LXI Output LAN Events menu. Key panels are removed as the number of possible LAN events decreases past a multiple of six, and the “More” keys are updated to reflect the new number of key panels in the LXI Output LAN Events menu. Events from the default list cannot be removed.

Remote Command:

Example:

Notes:

State Saved:

Range:

Initial S/W Revision:

:LXI:EVENt[:OUTPut]:LAN:REMove[:EVENt] “LANEVENT”

:LXI:EVEN:LAN:REM “LANEVENT”

The maximum length of the string is 16 characters.

Longer strings are concatenated and the resulting LAN Event is removed from the LXI Output LAN Event list.

Nothing happens if the LAN event was not introduced using the Add command.

No

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

226

System Functions

System

Remove All (Remote Command Only)

Clears the list of custom LAN events (those introduced using the Add command) that are available to output as a response to instrument events. As new LAN events are removed, keys are removed from the LXI Output LAN

Events menu. Key panels are removed as the number of possible LAN events decreases past a multiple of six, and the “More” keys are updated to reflect the new number of key panels in the LXI Output LAN Events menu.

Remote Command:

Example:

Notes:

:LXI:EVENt[:OUTPut]:LAN:REMove:ALL

:LXI:EVEN:LAN:REM:ALL

Only LAN Events added with the Add command are removed.

Default events cannot be removed.

Prior to A.02.00

Initial S/W Revision:

Source

Sets the instrument event that this LXI Output LAN event is tied to.

The possible instrument events are “WaitingForTrigger”, “Sweeping”, “Measuring”, “OperationComplete”, and

“Recalling”.

The key is labeled with the value of the selected source.

For the instrument event specific LXI Output LAN Events “WaitingForTrigger,” “Sweeping,” “Measuring,”

“OperationComplete,” and “Recalling,” this parameter is set to the corresponding source value and cannot be changed. For these events, the Source key does not appear.

WaitingForTrigger, Measuring, and Sweeping correspond to the standard trigger state machine activities for which they are named.

OperationComplete is low when a measurement operation is underway. For example, OperationComplete is low throughout a list sweep measurement, even though Sweeping, Measuring, and WaitingForTrigger will undergo a number of transitions. In this case, OperationComplete goes high when the entire list sweep is finished.

Recalling is high while the instrument is actively recalling a state.

Additionally, the Source parameter can be set to the name of any Input LAN Event. This causes the Output LAN

Event to be sent upon receipt of the named Input LAN Event. There is no front panel support for these events.

The default list of available Input LAN Events is:

• “LAN0”

• “LAN1”

• “LAN2”

• “LAN3”

• “LAN4”

• “LAN5”

• “LAN6”

• “LAN7”

Key Path

System, I/O Config, LXI, LXI Output LAN Events, LAN[n]

227

System Functions

System

Remote Command

Example

Notes

Preset

Preset

State Saved

Range

:LXI:EVENt[:OUTPut]:LAN[:SET]:SOURce “LANEVENT”,

“SourceEvent”

:LXI:EVENt[:OUTPut]:LAN[:SET]:SOURce? “LANEVENT”

:LXI:EVEN:LAN:SOUR “LANEVENT”,“WaitingForTrigger”

The maximum length of the string is 45 characters.

”Sweeping” (The Output LAN Events “WaitingForTrigger”, “Sweeping”,

“Measuring”, “OperationComplete”, and “Recalling” all have default source parameters that match their names)

Not affected by a Preset. The default values can be restored by pressing

Restore Defs, Input/Output Settings.

Preset/Default values:Sweeping” (The Output LAN Events

“WaitingForTrigger”, “Sweeping”, “Measuring”, “OperationComplete”, and

“Recalling” all have default source parameters that match their names)

Saved in instrument state.

“WaitingForTrigger”|“Sweeping”|“Measuring”|“OperationComplete”|“Recall ing”|”LAN0”|”LAN1”| ”LAN2”| ”LAN3”| ”LAN4”| ”LAN5”| ”LAN6”|

”LAN7”| any user-added Input LAN Event

Prior to A.02.00

Initial S/W Revision

Destination (Remote Command Only)

Outgoing LAN events are sent to the hosts enumerated in the destination expression. This expression takes the form of “host1:port1, host2:port2, ...” where port numbers are optional, and default to the IANA assigned TCP port

(5044). To designate a UDP broadcast at the default port, set the destination string to “” or “ALL”. To designate a

UDP broadcast at a specific port, set the destination string to “:port” or “ALL:port”.

Examples:

• “192.168.0.1:23”

• “agilent.com, soco.agilent.com”

• “agilent.com:80, 192.168.0.1”

Remote Command:

Example:

Notes:

Preset:

Preset:

State Saved:

:LXI:EVENt[:OUTPut]:LAN[:SET]:DESTination

“LANEVENT”,”destinationExpression”

:LXI:EVENt[:OUTPut]:LAN[:SET]:DESTination? “LANEVENT”

:LXI:EVEN:LAN:DEST “LANEVENT”,“host1, 192.168.0.1:80”

The maximum length of the string is 45 characters.

“ALL”

Not affected by a Preset. The default value of "ALL" can be restored by using the command:

:SYSTem:DEFault INPut

Saved in instrument state.

228

System Functions

System

Range:

Initial S/W Revision:

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

Drive

Determines the behavior of an output event.

• Normal designates typical operation, where both edges of the instrument event are transmitted,

• Off disables the LAN event.

• Wired-OR causes only one edge to be transmitted.

Key Path

Remote Command

Example

Preset

Preset

State Saved

Range

Initial S/W Revision

System, I/O Config, LXI, LXI Output LAN Events, LAN[n]

:LXI:EVENt[:OUTPut]:LAN[:SET]:DRIVe “LANEVENT”,

OFF|NORMal|WOR

:LXI:EVENt[:OUTPut]:LAN[:SET]:DRIVe? “LANEVENT”

:LXI:EVEN:LAN:DRIV “LANEVENT”,WOR

NORMal

Not affected by a Preset. The default value of "NORMal" can be restored by using the command:

:SYSTem:DEFault INPut

Saved in instrument state.

OFF|NORMal|WOR

Prior to A.02.00

Slope

Determines which instrument event transition results in a LAN packet being sent and whether or not that edge is inverted.

When the Drive parameter is set to Normal, a Slope of Negative causes both edges to be inverted before they are transmitted. A Positive Slope transmits the edges unaltered.

When the Drive parameter is set to WOR, only Positive edges are transmitted. When the Slope is Negative, a falling edge is inverted and sent as a rising edge. When the Slope is Positive, a rising edge is sent normally.

The following table illustrates the effects of the Slope and Drive parameters.

Slope Parameter Drive Parameter Action

1

1

0

0

Instrument Event

Edge

Negative

Positive

Negative

Positive

Off

Off

Off

Off

Not sent

Not sent

Not sent

Not sent

229

System Functions

System

1

1

0

0

1

1

0

0

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Positive

Normal

Normal

Normal

Normal

Wired OR

Wired OR

Wired OR

Wired OR

0

1

1

0

1

Not sent

Not sent

0

Key Path

Remote Command

Example

Preset

Preset

State Saved

Range

Initial S/W Revision

System, I/O Config, LXI, LXI Output LAN Events, LAN[n]

:LXI:EVENt[:OUTPut]:LAN[:SET]:SLOPe “LANEVENT”,

POSitive|NEGative

:LXI:EVENt[:OUTPut]:LAN[:SET]:SLOPe? “LANEVENT”

:LXI:EVEN:LAN:SLOP “LANEVENT”,POS

POSitive

Not affected by a Preset. The default value of "Positive" can be restored by using the command:

:SYSTem:DEFault INPut

Saved in instrument state.

POSitive|NEGative

Prior to A.02.00

Timestamp Delta

This parameter represents a time in seconds to add to the timestamp of the Output LAN Event. This timestamp delta allows the receiving instrument to delay its response until the time specified in the timestamp.

Key Path

Remote Command

Example

Preset

Preset

State Saved

System, I/O Config, LXI, LXI Output LAN Events, LAN[n]

:LXI:EVENt[:OUTPut]:LAN[:SET]:TSDelta “LANEVENT”,

<seconds>

:LXI:EVENt[:OUTPut]:LAN[:SET]:TSDelta? “LANEVENT”

:LXI:EVEN:LAN:TSD “LANEVENT”,10.5 s

0.0 s

Not affected by a Preset. The default value of "0.0 s" can be restored by using the command:

:SYSTem:DEFault INPut

Saved in instrument state.

230

System Functions

System

Range

Initial S/W Revision

0.0 – 1.7976931348623157 x 10308 s(Max Double)

Prior to A.02.00

Enabled

If this parameter is set to ON, this LAN Event is sent when the selected Source instrument event occurs.

Otherwise, this LAN Event is never output.

Key Path

Remote Command

Example

Preset

Preset

State Saved

Range

Initial S/W Revision

System, I/O Config, LXI, LXI Output LAN Events, LAN[n]

:LXI:EVENt[:OUTPut]:LAN[:SET]:ENABled

“LANEVENT”,ON|OFF|1|0

:LXI:EVENt[:OUTPut]:LAN[:SET]:ENABled? “LANEVENT”

:LXI:EVEN:LAN:ENAB “LAN0”,ON

OFF

Not affected by a Preset. The default value of "OFF" can be restored by using the command:

:SYSTem:DEFault INPut

Saved in instrument state.

OFF|ON|0|1

Prior to A.02.00

Count (Remote Command Only)

Returns the number of items in the LXI Output LAN Event List.

Remote Command:

Example:

Initial S/W Revision:

:LXI:EVENt[:OUTPut]:LAN:COUNt?

:LXI:EVEN:LAN:COUN?

Prior to A.02.00

Configure (Remote Command Only)

Allows the configuration of some parameters from a single SCPI command.

Remote Command:

Example:

Initial S/W Revision:

:LXI:EVENt[:OUTPut]:LAN[:SET]:CONFigure

“lanEvent”,<enabled>,<source>,<slope>,<drive>,<destinat ion>

:LXI:EVEN:LAN:CONF

“LAN0”,1,”WaitingForTrigger”,POS,NORM,”ALL”

Prior to A.02.00

Send (Remote Command Only)

Forces the instrument to send the requested LAN Event. The LAN Event must be enabled, otherwise this command

231

System Functions

System

is ignored.

Remote Command:

Example:

Initial S/W Revision:

:LXI:EVENt[:OUTPut]:LAN:SEND “LANEVENT”", RISE|FALL

:LXI:EVEN:LAN:SEND “LANEVENT”, FALL

Prior to A.02.00

Identifier (Remote Command Only)

Sets the string that will be placed in the peer-to-peer packet when the Output LAN Event is transmitted. The

Identifier is variable to allow for easier system debugging. The Identifier must be unique, for example the “LAN0” and “LAN1” output events cannot have identical identifiers.

Remote Command:

Example:

Notes:

State Saved:

Range:

Initial S/W Revision:

:LXI:EVENt[:OUTPut]:LAN[:SET]:IDENtifier “LANEVENT”,

“identifier”

:LXI:EVENt[:OUTPut]:LAN[:SET]:IDENtifier? "LANEVENT"

:LXI:EVEN:LAN:IDEN"LAN0","debugstring"

The maximum length of the string is 16 characters.

Nothing happens if the LAN event does not exist.

The default value is that the identifier is equivalent to the name of the LAN

Event.

Saved in instrument state.

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

IEEE 1588 Time (Remote Command Only)

Epoch Time (Remote Command Only)

If the device is selected as the IEEE 1588 master clock, this sets the clock using the number of seconds elapsed since January, 1 1970 at 00:00:00 in International Atomic Time (TAI). Epoch time is time zone invariant. Otherwise, this allows the user to query the epoch time.

Remote Command:

Example:

Notes:

Preset:

:LXI:CLOCk[:TIME][:VALue] <seconds>,<fractionalSeconds>

:LXI:CLOC 10020304.0 s,0.123456 s

The seconds argument must only contain values representing whole seconds.

For example 1243.0 s is acceptable, but 1243.01 results in an error.

Ignored when the device is not selected as the IEEE 1588 master clock.

The fractional portion is only accurate to the microseconds position.

Error generated if the seconds argument contains a fractional portion.

System time

232

System Functions

System

Preset:

State Saved:

Range:

Initial S/W Revision:

Not affected by a Preset. The default value of "System Time" can be restored by using the command:

:SYSTem:DEFault INPut

No

Seconds: 0.0 – 1.7976931348623157 x 10308 s (Max Double)

Fraction: 0.0 s – 0.999999 s

Prior to A.02.00

Remote Command:

Example:

Notes:

Preset:

State Saved:

Range:

:LXI:CLOCk[:TIME][:VALue]?

:LXI:CLOC?

The seconds argument must only contain values representing whole seconds.

For example 1243.0 s is acceptable, but 1243.01 results in an error.

Ignored when the device is not selected as the IEEE 1588 master clock.

The fractional portion is only accurate to the microseconds position.

Error generated if the seconds argument contains a fractional portion.

System time

No

Seconds: 0.0 – 1.7976931348623157 x 10308 s (Max Double)

Fraction: 0.0 s – 0.999999 s

Prior to A.02.00

Initial S/W Revision:

Seconds (Remote Command Only)

If the device is selected as the IEEE 1588 master clock, this sets the seconds portion of the clock.

Otherwise, this allows the user to query the seconds portion of the epoch time. Valid values are in discrete increments of whole seconds.

Remote Command:

Example:

Notes:

Preset:

Preset:

:LXI:CLOCk[:TIME]:SEConds <seconds>

:LXI:CLOCk[:TIME]:SEConds?

:LXI:CLOC:SEC 10020304.0

Ignored when the device is not selected as the IEEE 1588 master clock.

Error generated if the argument contains a fractional portion. For example

1243.0 s is acceptable, but 1243.01 results in an error.

System time

Not affected by a Preset. The default value of "System Time" can be restored by using the command:

:SYSTem:DEFault INPut

233

System Functions

System

State Saved:

Range:

Initial S/W Revision:

No

0.0 – 1.7976931348623157 x 10308 s (Max Double)

Prior to A.02.00

Fraction (Remote Command Only)

If the device is selected as the IEEE 1588 master clock, this sets the sub-second value of the clock.

Otherwise, this allows the user to query the sub-second value of the epoch time.

Remote Command:

Example:

Notes:

Preset:

State Saved:

Range:

Initial S/W Revision:

:LXI:CLOCk[:TIME]:FRACtion <fraction>

:LXI:CLOCk[:TIME]:FRACtion?

:LXI:CLOC:FRAC 10 ms

Ignored when the device is not selected as the IEEE 1588 master clock.

Only accurate to the microseconds position.

Sub-second value of system time

No

[0.0,1.0)

Prior to A.02.00

Local Time (Remote Command Only)

Returns the current local time formatted as a date time string.

Remote Command:

Example:

Notes:

Example:

Range:

Initial S/W Revision:

:LXI:CLOCk[:TIME]:LOCal?

:LXI:CLOC:LOC? ! Returns “5/15/2007 6:23:34.123456”

LXI:CLOCk[:TIME]:LOCal? Returns Any string constituting a valid date and time

Prior to A.02.00

Initial S/W Revision:

Leap Second Offset (Remote Command Only)

Enables you to set the leap second offset between the UTC and TAI time standards.

Remote Command:

:LXI:CLOCk[:TIME]:LSOFfset <integer>

:LXI:CLOCk[:TIME]:LSOFfset?

:LXI:CLOC:LSOF 55

0 – 2147483647 (Max Integer)

Prior to A.02.00

234

System Functions

System

International Atomic Time (Remote Command Only)

Retrieves the current time using the TAI format.

Remote Command:

Example:

Notes:

Initial S/W Revision:

:LXI:CLOCk[:TIME]:TAI?

:LXI:CLOC:TAI? ! “5/15/2007 6:23:34.123456”

:LXI:CLOCk[:TIME]:TAI? Returns Any string constituting a valid date and time

Prior to A.02.00

Time Zone (Remote Command Only)

Retrieves the current local time zone as an offset in hours, minutes, and seconds from Greenwich Mean

Time.

Remote Command:

Example:

Notes:

Initial S/W Revision:

:LXI:CLOCk[:TIME]:TZONe?

:LXI:CLOC:TZON?

:LXI:CLOC:TZON? returns "01:00:00" if the current local time zone is 1 hour ahead from Greenwich Mean Time

Prior to A.02.00

Daylight Savings (Remote Command Only)

Retrieves the current status of the Windows System setting for Daylight Savings Time. Whether or not daylight savings time is in effect influences the time zone parameter.

Remote Command:

Example:

Notes:

Preset:

Initial S/W Revision:

:LXI:CLOCk[:TIME]:DLSavings?

:LXI:CLOC:DLS?

:LXI:CLOC:DLS? Returns 1 when Daylight Savings Time is On and 0 if the when Daylight Savings Time is Off

The Windows system Daylight Savings setting.

Prior to A.02.00

Coordinated Universal Time (Remote Command Only)

Retrieves the current time using the UTC format.

Remote Command:

Example:

Notes:

Initial S/W Revision:

:LXI:CLOCk[:TIME]:UTC?

:LXI:CLOC:UTC? !“5/15/2007 6:23:34.123456”

:LXI:CLOC:UTC? Returns Any string constituting a valid date and time

Prior to A.02.00

235

System Functions

System

Time Marker (Remote Command Only)

Records the PTP time as a marker that can later be measured against the current PTP time. Typical use is to time the length of a sequence of instrument operations. There are 9 available markers with indices 1 –

9.

Remote Command:

Example:

Initial S/W Revision:

:LXI:CLOCk[:TIME]:MARKer[1]|2|3|4|5|6|7|8|9[:SET]

:LXI:CLOC:MARK1

:LXI:CLOC:MARK2

:LXI:CLOC:MARK3

:LXI:CLOC:MARK4

:LXI:CLOC:MARK5

:LXI:CLOC:MARK6

:LXI:CLOC:MARK7

:LXI:CLOC:MARK8

:LXI:CLOC:MARK9

:LXI:CLOC:MARK

Prior to A.02.00

Time Marker Clear (Remote Command Only)

Clears the recorded PTP time marker used to measure against the current PTP time. There are 9 available markers with indices 1 – 9.

Remote Command:

Example:

:LXI:CLOCk[:TIME]:MARKer[1]|2|3|4|5|6|7|8|9:CLEAr

:LXI:CLOC:MARK1:CLEA

:LXI:CLOC:MARK2:CLEA

:LXI:CLOC:MARK3:CLEA

:LXI:CLOC:MARK4:CLEA

:LXI:CLOC:MARK5:CLEA

:LXI:CLOC:MARK6:CLEA

:LXI:CLOC:MARK7:CLEA

:LXI:CLOC:MARK8:CLEA

:LXI:CLOC:MARK9:CLEA

:LXI:CLOC:MARK:CLEA

Prior to A.02.00

Initial S/W Revision:

Time Marker Delta (Remote Command Only)

Calculates and returns the delta time from the marker to the present PTP time. Also returns the seconds

236

System Functions

System

and sub-seconds portions of the start and end times. There are 9 available markers with indices 1 – 9.

Remote Command:

Example:

Notes:

Range:

Initial S/W Revision:

:LXI:CLOCk[:TIME]:MARKer[1]|2|3|4|5|6|7|8|9:DELTa?

:LXI:CLOC:MARK1:DELT? ! returns

<deltaTime>,<startSeconds>,<startFractionalSeconds>,<endSeconds>,<endF ractionalSeconds>

:LXI:CLOC:MARK2:DELT?

:LXI:CLOC:MARK3:DELT?

:LXI:CLOC:MARK4:DELT?

:LXI:CLOC:MARK5:DELT?

:LXI:CLOC:MARK6:DELT?

:LXI:CLOC:MARK7:DELT?

:LXI:CLOC:MARK8:DELT?

:LXI:CLOC:MARK9:DELT?

:LXI:CLOC:MARK:DELT?

:LXI:CLOCk[:TIME]:MARKer[1]|2|3|4|5|6|7|8|9:DELTa? Returns a value between 0.0 – 1.7976931348623157 x 10308 s (Max Double)

0.0 – 1.7976931348623157 x 10308 s (Max Double)

Prior to A.02.00

Measurement Data Timestamp (Remote Command Only)

Returns the beginning and ending times of the last measurement cycle. This command also returns the duration of the measurement cycle. These values correspond to the last rising and falling transition of the

Measuring instrument event.

Remote Command:

Example:

Notes:

Initial S/W Revision:

:LXI:CLOCk[:TIME]:MEASure[:DELTa]?

:LXI:CLOC:MEAS? !Returns 2.0,1145902.0,0.123456, ,1145904.0,0.123456

:LXI:CLOCk[:TIME]:MARKer[1]|2|3|4|5|6|7|8|9:DELTa? Returns a value between 0.0 – 1.7976931348623157 x 10308 s (Max Double)

Prior to A.02.00

Clear Measurement Data Timestamp (Remote Command Only)

Forces the return values of the Measurement Data Timestamp to zero until the next measurement cycle occurs. This command need not be issued for the Measurement Data Timestamp to be refreshed.

Remote Command:

Example:

Initial S/W Revision:

:LXI:CLOCk[:TIME]:MEASure:CLEar

:LXI:CLOC:MEAS:CLE

Prior to A.02.00

237

System Functions

System

Precision Time Protocol

Precision Time Protocol, as defined by IEEE 1588, is a method for synchronizing the time across a network. Instruments participating in the PTP network can coordinate activities using this common time base.

Accuracy (Remote Command Only)

Sets the typical offset from the correct time that a user can expect from the instrument PTP clock. This parameter is used when the instrument is selected as the Master clock. It should be set along with the time when configuring a master clock.

The value should be chosen by judging how precisely the clock can be set to the exact TAI time and the accuracy and drift of the clock’s underlying oscillator.

This is an input to the IEEE 1588 Best Master Clock algorithm.

Remote Command:

Example:

Preset:

Preset:

Range:

Example:

Preset:

Preset:

Range:

:LXI:CLOCk:PTP:ACCuracy

NS25|NS100|NS250|NS1000|NS2500|US10|US25|US100|US250|US

1000|US2500|MS10|MS25|MS100|MS1000|S10|GT10S|UNKNown

:LXI:CLOCk:PTP:ACCuracy?

:LXI:CLOC:PTP:ACC US25

GT10S

Not affected by a Preset. The default value of "GT10S" can be restored by using the command:

SYSTem:DEFault INPut

NS25|NS100|NS250|NS1000|NS2500|US10|US25|US100|US250|US1000|US

2500|MS10|MS25|MS100|MS1000|S10|GT10S|UNKNown

Prior to A.02.00

Initial S/W Revision:

Announce Interval (Remote Command Only)

Sets the time in seconds between PTP announce packets. A shorter interval makes the system more responsive to changes in the master clock at the cost of network bandwidth and packet processing time. The announce interval should be constant across all the instruments in the network. The announce interval will be rounded to the nearest non-negative integer power of two, with a maximum value of 16.

Remote Command:

:LXI:CLOCk:PTP:ANNounce:INTerval <interval>

:LXI:CLOCk:PTP:ANNounce:INTerval?

:LXI:CLOC:PTP:ANN:INT 1

2

Not affected by a Preset. The default value of "4" can be restored by using the command:

SYSTem:DEFault INPut

1|2|4|8|16

238

System Functions

System

Initial S/W Revision: Prior to A.02.00

Announce Receipt Time Out (Remote Command Only)

Sets the number of announce intervals that the instrument waits to receive an announce packet while in the Slave or

Listening. After this number of announce intervals, the instrument will transition to the Master state.

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:PTP:ANNounce:RTOut <numberOfIntervals>

:LXI:CLOCk:PTP:ANNounce:RTOut?

:LXI:CLOC:PTP:ANN:RTO 5

3

Not affected by a Preset. The default value of "3" can be restored by using the command:

SYSTem:DEFault INPut

2

10

Prior to A.02.00

Clock Class (Remote Command Only)

Returns a ranking of the master clock suitability relative to other clocks on the network. A lower value represents a more suitable clock.

Suitability is defined by the IEEE 1588 standard section 7.6.2.4

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:PTP:CCLass?

:LXI:CLOC:PTP:CCL?

248

Not affected by a Preset. The default value of "248" can be restored by using the command:

SYSTem:DEFault INPut

6

248

Prior to A.02.00

Deviation (Remote Command Only)

Returns the standard deviation of the instrument’s PTP time from the Grandmaster’s PTP time.

Remote Command:

Example:

:LXI:CLOCk:PTP:DEViation?

:LXI:CLOC:PTP:DEV?

239

System Functions

System

Initial S/W Revision: Prior to A.02.00

Domain (Remote Command Only)

The instrument synchronizes its clock only with other clocks in the same domain.

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:PTP:DOMain <domainNumber>

:LXI:CLOCk:PTP:DOMain?

:LXI:CLOC:PTP:DOM 0

0

Not affected by a Preset. The default value of "0" can be restored by using the command:

SYSTem:DEFault INPut

0

127

Prior to A.02.00

Offset (Remote Command Only)

Returns the difference between the instrument clock PTP time and the Master clock PTP time.

Remote Command:

Example:

Range:

Initial S/W Revision:

:LXI:CLOCk:PTP:OFFSet?

:LXI:CLOC:PTP:OFFS?

0.0 to – 1.7976931348623157 x 10308 s (Min Double)

Prior to A.02.00

First Priority (Remote Command Only)

Setting this parameter overrides the IEEE 1588 Best Master Clock algorithm. If an instrument’s First Priority parameter is smaller than all other clocks in its domain, it is chosen as the Master clock.

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

:LXI:CLOCk:PTP:PRIority:FIRSt <priority>

:LXI:CLOCk:PTP:PRIority:FIRSt?

:LXI:CLOC:PTP:PRI:FIRS 50

128

Not affected by a Preset. The default value of "128" can be restored by using the command:

SYSTem:DEFault INPut

0

255

240

System Functions

System

Initial S/W Revision: Prior to A.02.00

Second Priority (Remote Command Only)

When two or more clocks are determined to be equally good by the Best Master Clock algorithm, the clock with the lowest Second Priority value is chosen to be the Master Clock.

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:PTP:PRIority:SECond <priority>

:LXI:CLOCk:PTP:PRIority:SECond?

:LXI:CLOC:PTP:PRI:SEC 50

128

Not affected by a Preset. The default value of "128" can be restored by using the command:

SYSTem:DEFault INPut

0

255

Prior to A.02.00

State (Remote Command Only)

Returns the current state of the instrument’s PTP clock as defined in the IEEE 1588 standard.

Remote Command:

Example:

Range:

:LXI:CLOCk:PTP:STATe?

:LXI:CLOC:PTP:STAT?

INITializing|FAULty|DISabled|LISTening|PREMaster|

MASTer|PASSive|UNCalibrated|SLAVe

Prior to A.02.00

Initial S/W Revision:

Traceability (Remote Command Only)

Returns the quality of the instrument’s PTP clock source of time when chosen as the Grand Master clock.

This parameter is used by the Best Master Clock algorithm.

Remote Command:

Example:

Range:

Initial S/W Revision:

:LXI:CLOCk:PTP:TRACeability?

:LXI:CLOC:PTP:TRAC?

ATOMic|GPS|RADio|PTP|NTP|HANDset|OTHer|OSCillator

Prior to A.02.00

241

System Functions

System

Variance (Remote Command Only)

Returns the variance of the instrument’s PTP clock time relative to the Master’s PTP clock time.

Remote Command:

Example:

Range:

Initial S/W Revision:

:LXI:CLOCk:PTP:VARiance?

:LXI:CLOC:PTP:VAR?

0.0 – 1.7976931348623157 x 10308 (Max Double)

Prior to A.02.00

Sync Interval (Remote Command Only)

Sets the rate at which PTP sync packets are transmitted when this instrument is acting as a Master PTP clock. The values must be integer powers of 2.

Remote Command:

Example:

Preset:

Preset:

Range:

Initial S/W Revision:

:LXI:CLOCk:PTP:SINTerval <seconds>

:LXI:CLOC:PTP:SINT 0.25s

1

Not affected by a Preset. The default value of "1" can be restored by using the command:

SYSTem:DEFault INPut

0.0625s|0.125s|0.25s|0.5s|1s|2s

Prior to A.02.00

Remote Command:

Example:

Preset:

Range:

Initial S/W Revision:

:LXI:CLOCk:PTP:SINTerval?

:LXI:CLOC:PTP:SINT?

1

0.0625s|0.125s|0.25s|0.5s|1s|2s

Prior to A.02.00

Delay Request Interval (Remote Command Only)

This property is used by the master clock to specify the interval between delay request packets sent from the slave to the master clock. Slaves use a randomly-chosen interval, with mean equal to this property.

The value for this parameter must be an integer power of two.

Remote Command:

Example:

Preset:

:LXI:CLOCk:PTP:DRINterval <seconds>

:LXI:CLOC:PTP:DRIN 15 ms

8 s

242

System Functions

System

Preset:

Range:

Initial S/W Revision:

Not affected by a Preset. The default value of "8 s" can be restored by using the command:

SYSTem:DEFault INPut

1 s | 2 s | 4 s | 8 s | 16 s | 32 s

Prior to A.02.00

Remote Command:

Example:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:PTP:DRINterval?

:LXI:CLOC:PTP:DRIN 15 ms

8 s

0.0 s

2^32 = 4294967296 s

Prior to A.02.00

Grand Master

In the IEEE 1588 the best clock in the system, as determined by the Best Master Clock algorithm, is chosen as the Grand Master clock. The Grand Master clock is the ultimate source of time which all other clocks attempt to synchronize to. If the network is limited to a single subnet, the Master clock and the

Grand Master clock are synonymous.

If a network spans multiple subnets, boundary clocks must be utilized to segment the network. Each subnet then selects a Master Clock using the Best Master Clock algorithm. From this population of

Master Clocks, the best clock is selected to be the Grand Master clock.

Accuracy (Remote Command Only)

Returns the relative accuracy of the Grand Master clock.

Remote Command:

Example:

Range:

:LXI:CLOCk:PTP:GMASter:ACCuracy?

:LXI:CLOC:PTP:GMAS:ACC? !For example, this might return GT10S.

25NS|100NS|250NS|1US|2.5US|10US|25US|100US|250US|1MS|2.5MS|10M

S|25MS|100MS|1S|10S|GT10S|UNKNown

Prior to A.02.00

Initial S/W Revision:

MAC Address (Remote Command Only)

Returns the Grand Master’s MAC Address.

Remote Command:

Example:

Range:

Initial S/W Revision:

:LXI:CLOCk:PTP:GMASter:MADDress?

:LXI:CLOC:PTP:GMAS:MADD? !For example, this might return

"00–00–50–1e-ca-ad".

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

243

System Functions

System

Traceability (Remote Command Only)

Describes the quality of the Grand Master PTP clock’s source of time.

Remote Command:

Example:

Range:

Initial S/W Revision:

:LXI:CLOCk:PTP:GMASter:TRACeability?

:LXI:CLOC:PTP:GMAS:TRAC? !For example, this might return OSC.

ATOMic|GPS|RADio|PTP|NTP|HANDset|OTHer|OSCillator

Prior to A.02.00

Master

Instruments on the same subnet use the Best Master Clock algorithm to select a Master clock. The

Master clock is used as the basis for synchronization.

MAC Address (Remote Command Only)

Returns the Master’s MAC Address.

Remote Command:

Example:

Range:

Initial S/W Revision:

:LXI:CLOCk:PTP:MASTer:MADDress?

:LXI:CLOC:PTP:MAST:MADD?

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

Servo Algorithm (Remote Command Only)

The Servo Algorithm parameters are considered advanced settings for tweaking IEEE 1588 performance.

Log (Remote Command Only)

The Servo Log records measurements of the offset between the instrument’s PTP clock and the Master’s PTP clock.

It also records the packet travel time for Master-to-Slave and Slave-to-Master transactions.

Next (Remote Command Only)

Retrieves and removes the oldest entry from the Servo Log. The format for a servo log entry is as follows

Sample Index:

Time Seconds:

Time Fraction:

Offset Seconds:

Average Delay Seconds:

Master Delay Seconds:

Slave Delay Seconds: integer representing entry order seconds portion of the entry timestamp sub-second portion of the entry timestamp offset between the instrument’s PTP clock and the Master’s PTP clock the average measured transmission delay

Master-to-Slave packet travel time

Slave-to-Master packet travel time

244

System Functions

System

Remote Command:

Example:

Range:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm:LOG[:NEXT]?

:LXI:CLOC:SALG:LOG?

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

Circular (Remote Command Only)

Sets the behavior for entries occurring while the Servo Log is full.

• If Circular is set to 1, incoming events overwrite the oldest events in the log.

• If Circular is set to 0, incoming events are discarded.

Remote Command:

Example:

Preset:

Preset:

:LXI:CLOCk:SALGorithm:LOG:CIRCular[:ENABled] ON|OFF|0|1

:LXI:CLOCk:SALGorithm:LOG:CIRCular[:ENABled]?

:LXI:CLOC:SALG:LOG:CIRC 1

1

Not affected by a Preset. The default value of "1" can be restored by using the command:

SYSTem:DEFault INPut

ON|OFF|0|1

Prior to A.02.00

Range:

Initial S/W Revision:

Beginning Entry (Remote Command Only)

Sets or freezes the beginning entry of the log when in circular mode to the most recently added entry at the time of the command. This is so that the :LXI:EVENt:LOG:ENTtry? command has a reference entry for indexing individual entries in the log.

Remote Command:

Example:

Initial S/W Revision:

Clear (Remote Command Only)

Clears all entries from the Servo Log.

:LXI:CLOCk:SALGorithm:LOG:CIRCular:FBENtry

LXI:CLOCk:SALG:LOG:CIRC:FBEN

Prior to A.02.00

Remote Command:

Example:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm:LOG:CLEar

:LXI:CLOC:SALG:LOG:CLE

Prior to A.02.00

245

System Functions

System

Count (Remote Command Only)

Returns the number of unread entries in the Servo Log.

Remote Command:

Example:

Range:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm:LOG:COUNt?

:LXI:CLOC:SALG:LOG:COUN?

0 - IEEE 1588 Servo Log Size

Prior to A.02.00

Enabled (Remote Command Only)

• When the Servo Log is disabled, no events are recorded.

• When it is enabled, the Servo Log is active.

Remote Command:

Example:

Preset:

Preset:

Range:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm:LOG:ENABled ON|OFF|0|1

:LXI:CLOCk:SALGorithm:LOG:ENABled?

:LXI:CLOC:SALG:LOG:ENAB 1

0

Not affected by a Preset. The default value of "0" can be restored by using the command:

SYSTem:DEFault INPut

ON|OFF|0|1

Prior to A.02.00

Size (Remote Command Only)

Sets the maximum number of entries to store in the Servo Log.

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm:LOG:SIZE <maxLogEntries>

:LXI:CLOCk:SALGorithm:LOG:SIZE?

:LXI:CLOC:SALG:LOG:SIZE 100

256

Not affected by a Preset. The default value of "256" can be restored by using the command:

SYSTem:DEFault INPut

0

1024

Prior to A.02.00

246

System Functions

System

All (Remote Command Only)

Non-destructively returns the entire contents of the Servo Log.

Remote Command:

Example:

Range:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm:LOG:ALL?

:LXI:CLOC:SALG:LOG?

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

Specific Entry (Remote Command Only)

Non-destructively returns a specifically indexed entry from within the Servo Log.

Remote Command:

Example:

Range:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm:LOG:ENTRy? <intIndex>

:LXI:CLOC:SALG:LOG? 0 !Returns the oldest entry in the Servo Log.

Example of result :

"1,1208978798,139644871,0.000000000,3.393600e+038,0.000000000,0.000

000000,0.000000000"

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

Statistics (Remote Command Only)

Returns the long-term statistics of the servo log that characterizes the performance of the instrument PTP clock’s offset from the master PTP clock. The statistics include the following values:

• Number of samples (an integer)

• Mean offset (a double)

• Standard deviation of the offset (a double)

• Maximum offset (a double)

• Minimum offset (a double)

Remote Command:

Example:

:LXI:CLOCk:SALGorithm:LOG:STATistics[:DATA]?

:LXI:CLOC:SALG:LOG:STAT? ! Example of result :

"3643,0.000000000,0.000000000,0.000000000,0.000000000"

Prior to A.02.00

Initial S/W Revision:

Clear Statistics (Remote Command Only)

Resets the long-term servo performance statistics.

Remote Command:

Example:

:LXI:CLOCk:SALGorithm:LOG:STATistics:CLEar

:LXI:CLOC:SALG:LOG:STAT? ! Example of result :

"3643,0.000000000,0.000000000,0.000000000,0.000000000"

247

System Functions

System

Initial S/W Revision: Prior to A.02.00

Asymmetry (Remote Command Only)

Sets the difference in seconds between the Master-to-Slave packet travel time and the Slave-to-Master packet travel time.

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm[:SET]:ASYMmetry <seconds>

:LXI:CLOCk:SALGorithm[:SET]:ASYMmetry?

:LXI:CLOC:SALG:ASYM 15 ns

0.0 s

Not affected by a Preset. The default value of "0.0 s " can be restored by using the command:

SYSTem:DEFault INPut

–1

1

Prior to A.02.00

Coarse/Fine Threshold (Remote Command Only)

Determines when the PTP clock Servo algorithm uses the ‘Fine’ or ‘Coarse’ parameters for adjusting the instrument’s PTP clock time. The threshold is measured against a running estimate of the servo variance.

Coarse mode causes a slave clock to converge with the master clock more quickly, but it is more sensitive to noise, while Fine mode filters out noise more effectively, but takes longer to converge.

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm[:SET]:CFTHreshold

<secondsSquared>

:LXI:CLOCk:SALGorithm[:SET]:CFTHreshold?

:LXI:CLOC:SALG:CFTH 0.25

1.0e–11

Not affected by a Preset. The default value of "1.0e–11 " can be restored by using the command:

SYSTem:DEFault INPut

0

1

Prior to A.02.00

Coarse Proportional Constant (Remote Command Only)

This constant is used by the servo when above the Coarse/Fine Threshold variance. Decreasing this constant causes the servo to become less responsive to both noise in the system and changes in the Master Clock’s rate. Conversely, increasing this constant causes the servo to respond more energetically to both system noise and changes in the

248

System Functions

System

Master Clock’s rate.

The ratio between the Proportional and Integral constants should remain roughly constant.

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm[:SET]:CPConstant <servoConstant>

:LXI:CLOCk:SALGorithm[:SET]:CPConstant?

:LXI:CLOC:SALG:CPC 0.5

0.4

Not affected by a Preset. The default value of "0.4 " can be restored by using the command:

SYSTem:DEFault INPut

0

1

Prior to A.02.00

Coarse Integral Constant (Remote Command Only)

This constant is used by the servo when above the Coarse/Fine Threshold variance. Decreasing this constant causes the servo to become less responsive to both noise in the system and changes in the Master Clock’s rate. Conversely, increasing this constant causes the servo to respond more energetically to both system noise and changes in the

Master Clock’s rate.

The ratio between the Proportional and Integral constants should remain roughly constant.

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm[:SET]:CIConstant <servoConstant>

:LXI:CLOCk:SALGorithm[:SET]:CIConstant?

:LXI:CLOC:SALG:CIC 0.5

0.2

Not affected by a Preset. The default value of "0.2 " can be restored by using the command:

SYSTem:DEFault INPut

0

1

Prior to A.02.00

Fine Proportional Constant (Remote Command Only)

This constant is used by the servo when below the Coarse/Fine Threshold variance. Decreasing this constant causes the servo to become less responsive to both noise in the system and changes in the Master Clock’s rate. Conversely, increasing this constant causes the servo to respond more energetically to both system noise and changes in the

Master Clock’s rate.

249

System Functions

System

The ratio between the Proportional and Integral constants should remain roughly constant.

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm[:SET]:FPConstant <servoConstant>

:LXI:CLOCk:SALGorithm[:SET]:FPConstant?

:LXI:CLOC:SALG:FPC 1

0.35

Not affected by a Preset. The default value of "0.35 " can be restored by using the command:

SYSTem:DEFault INPut

0

1

Prior to A.02.00

Fine Integral Constant (Remote Command Only)

This constant is used by the servo when below the Coarse/Fine Threshold variance. Decreasing this constant causes the servo to become less responsive to both noise in the system and changes in the Master Clock’s rate. Conversely, increasing this constant causes the servo to respond more energetically to both system noise and changes in the

Master Clock’s rate.

The ratio between the Proportional and Integral constants should remain roughly constant.

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm[:SET]:FIConstant <servoConstant>

:LXI:CLOCk:SALGorithm[:SET]:FIConstant?

:LXI:CLOC:SALG:FIC 0.6

0.05

Not affected by a Preset. The default value of "0.05" can be restored by using the command:

SYSTem:DEFault INPut

0

1

Prior to A.02.00

Maximum Outlier Discard Count (Remote Command Only)

Sets the maximum number of outlier packets to ignore. After this maximum is exceeded, the next packet is accepted, regardless of whether or not it is flagged as an outlier.

Remote Command:

:LXI:CLOCk:SALGorithm[:SET]:OMAXimum

<consecutiveSamples>

:LXI:CLOCk:SALGorithm[:SET]:OMAXimum?

250

System Functions

System

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOC:SALG:OMAX 3

5

Not affected by a Preset. The default value of "5" can be restored by using the command:

SYSTem:DEFault INPut

0

25

Prior to A.02.00

Outlier Threshold (Remote Command Only)

Defines the threshold for determining whether a packet is considered a statistical outlier. If a sync or delay request is held up in a switch for a significant amount of time, the quality of synchronization will be perturbed. The servo ignores anything outside the outlier threshold. This parameter is expressed as a number of standard deviations from the currently measured average packet latency. Note that the value can be set to fractional standard deviations.

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm[:SET]:OTHReshold

<standardDeviations>

:LXI:CLOCk:SALGorithm[:SET]:OTHReshold?

:LXI:CLOC:SALG:OTHR 1.0

5.0

Not affected by a Preset. The default value of "5.0" can be restored by using the command:

SYSTem:DEFault INPut

0.25

6.0

Prior to A.02.00

Outlier Threshold Enable (Remote Command Only)

Enables the outlier threshold to determine whether or not outliers are discarded.

Remote Command:

Example:

Preset:

Preset:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm[:SET]:OTENable ON|OFF|1|0

:LXI:CLOCk:SALGorithm[:SET]:OTENable?

:LXI:CLOC:SALG:OTEN OFF

OFF

Not affected by a Preset. The default value of "OFF" can be restored by using the command:

SYSTem:DEFault INPut

Prior to A.02.00

251

System Functions

System

Set/Steer Threshold (Remote Command Only)

If the instrument’s clock deviates from the master by an amount equal to or greater than this threshold, it is reset to match the master rather than being gradually steered toward it.

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm[:SET]:STHReshold <seconds>

:LXI:CLOCk:SALGorithm[:SET]:STHReshold?

:LXI:CLOC:SALG:STHR 15 ms

0.1 s

Not affected by a Preset. The default value of "0.1 s " can be restored by using the command:

SYSTem:DEFault INPut

0.0001

10.0

Prior to A.02.00

Configure (Remote Command Only)

Allows the configuration of some of the above parameters from a single SCPI command.

Remote Command:

Example:

Initial S/W Revision:

:LXI:CLOCk:SALGorithm[:SET]:CONFigure <asymmetry>,

<coarse fine threshold>, <cpc>, <cic>, <fpc>, <fic>,

<maximum outlier discard>, <outlier threshold>,

<set/steer threshold>

:LXI:CLOC:SALG:CONF 0.0, 2.0E–13, 0.4, 0.2, 0.35, 0.05, 5, 2.0E–4, 0.1s

Prior to A.02.00

Synchronization (Remote Command Only)

Synchronization parameters are used to control the behavior of the LAN communication used to achieve converging times across the system.

Master (Remote Command Only)

Reports whether or not the device has been selected as the PTP master clock.

Remote Command:

Example:

Range:

Initial S/W Revision:

:LXI:CLOCk:SYNC:MASTer?

:LXI:CLOC:SYNC:MAST?

ON|OFF|0|1

Prior to A.02.00

252

System Functions

System

Local Enabled (Remote Command Only)

Enable steering of the local clock with the PTP IEEE 1588 clock.

Remote Command:

Example:

Preset:

Preset:

Range:

Initial S/W Revision:

:LXI:CLOCk:SYNC:LOCal:ENABled ON|OFF|0|1

:LXI:CLOCk:SYNC:LOCal:ENABled?

:LXI:CLOC:SYNC:LOC:ENAB ON

OFF

Not affected by a Preset. The default value of "ON" can be restored by using the command:

SYSTem:DEFault INPut

ON|OFF|0|1

Prior to A.02.00

Local Interval (Remote Command Only)

The local clock is updated after the time set in the Local Interval elapses.

Remote Command:

Example:

Preset:

Preset:

Min:

Max:

Initial S/W Revision:

:LXI:CLOCk:SYNC:LOCal:INTerval

:LXI:CLOCk:SYNC:LOCal:INTerval?

:LXI:CLOC:SYNC:LOC:INT 60

60

Not affected by a Preset. The default value of "60 " can be restored by using the command:

SYSTem:DEFault INPut

0

3600

Prior to A.02.00

Instrument Status Events

Instrument status events represent internal state changes that are important for synchronization. The recognized Status Events are:

WaitingForTrigger:

Sweeping:

Transitions high when the measurement is awaiting a trigger event.

Transitions low when the trigger event occurs or the measurement is aborted.

Transitions high when the instrument begins sweeping. Transitions low when the sweep is completed.

low when measurement data is no longer being recorded.

253

System Functions

System

OperationComplete:

Recalling:

Transitions low when a measurement operation is in progress. In some cases (for example, ListSweep), multiple sweeps are taken during a single measurement operation. Transitions high when the measurement operation is finished or aborted.

Transitions high while the instrument is recalling a previous state.

Transitions low when a recall is not in progress.

Enable (Remote Command Only)

Setting the enabled parameter to ON enables the selected instrument event to be used as a source for

Output LAN Events. Enabling an Instrument Status Event also causes the event to appear in the Event

Log.

Remote Command:

Example:

Preset:

Preset:

State Saved:

Range:

Initial S/W Revision:

:LXI:EVENt:STATus[:ENABled] “STATUSEVENT”,ON|OFF|1|0

:LXI:EVEN:STAT “WaitingForTrigger”,1

ON

Not affected by a Preset. The default value of "1" can be restored by using the command, :SYSTem:DEFault INPut.

Saved in instrument state.

1|0|ON|OFF

Prior to A.02.00

Remote Command:

Example:

Preset:

State Saved:

Range:

Initial S/W Revision:

:LXI:EVENt:STATus[:ENABled]? “STATUSEVENT”

:LXI:EVEN:STAT? “WaitingForTrigger”

Returns 1 if previously enabled. Otherwise, returns 0.

ON

Saved in instrument state.

1|0|ON|OFF

Prior to A.02.00

LXI State Recall

The LXI Event system can be used to recall states upon receipt of LXI Input LAN Events.

Location (Remote Command Only)

This parameter is used to store the file paths of the state files to be recalled when each Input LAN Event is received. Since each LAN Event has its own Location entry, a given state is capable of branching to at least 8 different states. If custom Input events are added, an even greater branching factor is possible.

When setting up state transitions, it is important to set the location of the next state before saving. This

254

System Functions

System

way, when the saved state is recalled, the next state locations are also automatically recalled.

Remote Command:

Example:

Notes:

State Saved:

Range:

Initial S/W Revision:

:LXI:EVENt:INPut:LAN:LOCation “LANEVENT”,”path”

:LXI:EVEN:INP:LAN:LOC “LANEVENT”,”c:\states\state01.state”

The maximum length of the string is 512 characters.

Saved in instrument state.

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

Remote Command:

Example:

Notes:

State Saved:

Range:

Initial S/W Revision:

:LXI:EVENt:INPut:LAN:LOCation? “LANEVENT”

:LXI:EVEN:INP:LAN:LOC? “LANEVENT”

Returns “c:\states\state01.state” if that value was previously entered

The maximum length of the string is 512 characters.

Saved in instrument state.

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

Disable All (Remote Command Only)

Causes all LXI Input LAN Events to go into the disabled state (Enabled = OFF).

Remote Command:

Example:

Initial S/W Revision:

:LXI:EVENt:INPut:LAN:DISable:ALL

:LXI:EVEN:INP:LAN:DIS:ALL

Prior to A.02.00

Add (Remote Command Only)

Adds the provided string to the list of possible LAN events to Input as a response to instrument events.

As new LAN events are added, keys are generated in the LXI Input LAN Events menu. New key panels are generated as the number of possible LAN events increases past a multiple of six, and the “More” keys are updated to reflect the new number of key panels in the LXI Input LAN Events menu.

Remote Command:

Example:

Notes:

State Saved:

Range:

:LXI:EVENt:INPut:LAN:ADD “LANEVENT”

:LXI:EVEN:INP:LAN:ADD “LANEVENT”

The maximum length of the string is 16 characters.

Longer strings are concatenated and added to the LXI Input LAN Event list.

No event is added if the LAN Event already exists.

No

Uppercase, Lowercase, Numeric, Symbol except for comma or semicolon

255

System Functions

System

Initial S/W Revision: Prior to A.02.00

Remove (Remote Command Only)

Removes the provided string from the list of LXI Input LAN Events. As new LAN events are removed, keys are removed from the LXI Input LAN Events menu. Key panels are removed as the number of possible LAN events decreases past a multiple of six, and the “More” keys are updated to reflect the new number of key panels in the LXI Input LAN Events menu. Events from the default list cannot be removed.

Remote Command:

Example:

Notes:

State Saved:

Range:

Initial S/W Revision:

:LXI:EVENt:INPut:LAN:REMove[:EVENt] “LANEVENT”

:LXI:EVEN:INP:LAN:REM “LANEVENT”

The maximum length of the string is 16 characters.

Longer strings are concatenated and the resulting LAN Event is removed from the LXI Input LAN Event list.

Nothing happens if the LAN event was not introduced using the Add command.

No

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

Remove All (Remote Command Only)

Clears the list of custom LAN events (those introduced using the Add command). As new LAN events are removed, keys are removed from the LXI Input LAN Events menu. Key panels are removed as the number of possible LAN events decreases past a multiple of six, and the “More” keys are updated to reflect the new number of key panels in the LXI Input LAN Events menu.

Remote Command:

Example:

Notes:

:LXI:EVENt:INPut:LAN:REMove:ALL

:LXI:EVEN:INP:LAN:REM:ALL

Only LAN Events added with the Add command are removed. Default events cannot be removed.

Prior to A.02.00

Initial S/W Revision:

Filter (Remote Command Only)

Only LXI Input LAN Events coming from hosts matching the filter string are processed. There is no Key

Path to this command

The syntax for specifying a filter is as follows:

Filter == ( [host[:port]] | [ALL[:port]] ) [,Filter]

Specifying an empty string means that LXI trigger packets are accepted as an Input from any port on any host on the network via either TCP or UDP.

256

System Functions

System

Specifying only the port means that any host communicating over that port can send events.

Specifying ALL indicates that UDP multicast packets are accepted if they are directed to the IANA assigned multicast address on the IANA assigned default port, or the designated port if specified.

Examples:

• “192.168.0.1:23”

• “agilent.com, soco.agilent.com”

• “agilent.com:80, 192.168.0.1”

• The LXI:EVENt:INPut:LAN:FILTer command applies only to state events and will have no effect on trigger events, even when both are tied to the same event name (like "LAN0"). Similarly, the

TRIGger:LXI:LAN:FILTer command applies only to trigger events and will have no effect on state events.

Remote Command:

Example:

Notes:

State Saved:

Range:

Initial S/W Revision:

:LXI:EVENt:INPut:LAN:FILTer “LANEVENT”,“filterString”

:LXI:EVENt:INPut:LAN:FILTer?

:LXI:EVEN:INP:LAN:FILT “LAN0”,“agilent.com”

:LXI:EVEN:INP:LAN:FILT?

The maximum length of the string is 45 characters. Nothing happens if the

LAN event does not exist.

Saved in instrument state.

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

Identifier (Remote Command Only)

Sets the string that is expected to arrive over the LAN for a given Input LAN Event to occur. The

Identifier is variable to allow for easier system debugging.

Remote Command:

Example:

Notes:

State Saved:

Range:

Initial S/W Revision:

:LXI:EVENt:INPut:LAN:IDENtifier “LANEVENT”,”identifier”

:LXI:EVENt:INPut:LAN:IDENtifier? “LANEVENT”

:LXI:EVEN:INP:LAN:IDEN “LAN0”,“debugstring”

The maximum length of the string is 16 characters.

Nothing happens if the LAN event does not exist.

The default value is that the identifier is equivalent to the name of the LAN

Event.

Saved in instrument state.

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

257

System Functions

System

Detection (Remote Command Only)

Selects the triggering option..

• Selecting “Rise” causes the instrument to trigger on the receipt of a signal low LAN Event followed by a signal high LAN Event.

• Selecting “Fall” causes the instrument to trigger on the receipt of a signal high LAN Event followed by a signal low LAN Event.

• Selecting “High” causes the instrument to trigger on every signal high LAN Event.

• Selecting “Low” causes the instrument to trigger on every signal low LAN Event.

Remote Command:

Example:

Notes:

Preset:

Preset:

State Saved:

Range:

Readback Text:

Initial S/W Revision:

:LXI:EVENt:INPut:LAN[:SET]:DETection “LANEVENT”,

HIGH|LOW|RISE|FALL

:LXI:EVENt:INP:LAN:DET “LANEVENT”,HIGH

If a non existent LAN event is passed in the lanEvent argument, the command is ignored

HIGH

Not affected by a Preset. The default value of "HIGH" can be restored by using the remote command:

:SYSTem:DEFault INPut

Saved in instrument state.

HIGH | LOW | RISE | FALL

Currently selected detection type

Prior to A.02.00

Remote Command:

Example:

Notes:

Preset:

State Saved:

Range:

Readback Text:

Initial S/W Revision:

:LXI:EVENt:INPut:LAN[:SET]:DETection? “LANEVENT”

:LXI:EVEN:INP:LAN:DET? “LANEVENT”

If a non existent LAN event is passed in the lanEvent argument, the command is ignored

HIGH

Saved in instrument state.

HIGH | LOW | RISE | FALL

Currently selected detection type

Prior to A.02.00

Enabled (Remote Command Only)

When the Enabled parameter is set to ON, receiving the given LAN Event causes the instrument to transition to the state held in the Next State Slot.

258

System Functions

System

When the Enabled parameter is OFF, the Input LAN Event is ignored.

Remote Command:

Example:

Preset:

Preset:

State Saved:

Range:

Initial S/W Revision:

:LXI:EVENt:INPut:LAN[:SET]:ENABled

“LANEVENT”,ON|OFF|1|0

:LXI:EVEN:INP:LAN:ENAB “LAN0”,1

OFF

Not affected by a Preset. The default value of "OFF" can be restored by using the remote command:

:SYSTem:DEFault INPut

Saved in instrument state.

1|0

Prior to A.02.00

Remote Command:

Example:

Preset:

State Saved:

Range:

Initial S/W Revision:

:LXI:EVENt:INPut:LAN[:SET]:ENABled? “LANEVENT”

:LXI:EVEN:INP:LAN:ENAB? “LAN0”

OFF

Saved in instrument state.

1|0

Prior to A.02.00

Count (Remote Command Only)

Returns the number of items in the LXI Input LAN Event List.

Remote Command:

Example:

Initial S/W Revision:

:LXI:EVENt:INPut:LAN:COUNt?

:LXI:EVEN:INP:LAN:COUN?

Prior to A.02.00

List (Remote Command Only)

Returns a list of all of the valid LXI Input LAN Event names.

Remote Command:

Example:

Preset:

State Saved:

Initial S/W Revision:

:LXI:EVENt:INPut:LAN:LIST?

:LXI:EVEN:INP:LAN:LIST?

Returns “LAN0”, “LAN1”, “LAN2”, “LAN3”, “LAN4”, “LAN5”, “LAN6”,

“LAN7”

“LAN0”, “LAN1”, “LAN2”, “LAN3”, “LAN4”, “LAN5”, “LAN6”, “LAN7”

Saved in instrument state.

Prior to A.02.00

259

System Functions

System

Configure (Remote Command Only)

Allows the configuration of some of the above parameters from a single SCPI command.

Remote Command:

Example:

Initial S/W Revision:

:LXI:EVENt:INPut:LAN[:SET]:CONFigure “lanEvent” ,

<enab>, <detection>, <filter>, <identifier>

:LXI:EVEN:INP:LAN:CONF "LAN0",1,FALL,"FILTER","DEBUG"

Prior to A.02.00

Restore Defaults

Provides incremental initialization of the system setting groups along with supporting a comprehensive reset of the entire instrument back to a factory default state. The menu selections are the groups of system settings and when one is selected, that particular group of system settings is reset back to their default values.

Key Path

Mode

Remote Command

Example

State Saved

Initial S/W Revision

System

All

:SYSTem:DEFault [ALL]|ALIGn|INPut|MISC|MODes|PON

SYST:DEF

No

Prior to A.02.00

Restore Input/Output Defaults

Causes the group of settings and data associated with Input/Output front-panel key to be a reset to their default values. This level of Restore System Defaults does not affect any other system settings, mode settings and does not cause a mode switch. .

Confirmation is required to restore the Input/Output setting. The confirmation dialog is:

Key Path

260

System, Restore System Defaults

System Functions

System

Example

Initial S/W Revision

:SYST:DEF INP

Prior to A.02.00

Restore Power On Defaults

This selection causes the Power On settings to be a reset to their default value. This level of Restore

System Defaults does not affect any other system settings, mode settings and does not cause a mode switch. The Power On settings and their default values are Power On Type reset to Mode and

Input/Output Defaults and Power On Application reset to whatever the factory set as its default value.

Confirmation is required to restore the factory default values. The confirmation dialog is:

Key Path

Example

Initial S/W Revision

System, Restore System Defaults

:SYST:DEF PON

Prior to A.02.00

Restore Align Defaults

This selection causes the Alignment system settings to be a reset to their default values. This does not affect any Alignment data stored in the system. This level of Restore System Defaults does not affect any other system settings, mode settings and does not cause a mode switch.

After performing this function, it may impact the auto-alignment time of the instrument until a new alignment baseline has been established.

Confirmation is required to restore the factory default values. The confirmation dialog is:

261

System Functions

System

Key Path

Example

Initial S/W Revision

System, Restore System Defaults

:SYST:DEF ALIG

Prior to A.02.00

Restore Misc Defaults

This selection causes miscellaneous system settings to be reset to their default values. With this reset, you lose the GPIB address and it is reset to 18, so this should be used with caution. This level of Restore

System Defaults does not affect any other system settings, mode settings and does not cause a mode switch. This miscellaneous group contains the rest of the settings that have not been part of the other

Restore System Defaults groups. The following table is a complete list of settings associated with this group:

Miscellaneous Setting

Verbose SCPI

GPIB Address

Auto File Name Number

Save Type

State Save To

Screen Save To

DISP:ENABle

Full Screen

SCPI Telnet

SCPI Socket

SICL Server

Display Intensity

Display Backlight

Display Theme

Off

ON

ON

ON

100

ON

TDColor

Default Value

Off

18

000

State

Register 1

SCREEN000.png

ON

262

System Functions

System

System Annotation

The SYST:PRES:TYPE

ON

MODE

Confirmation is required to restore the factory default values. The confirmation dialog is:

Key Path

Example

Initial S/W Revision

System, Restore System Defaults

:SYST:DEF MISC

Prior to A.02.00

Restore Mode Defaults (All Modes)

This selection resets all of the modes in the instrument back to their default state just as a Restore Mode

Defaults does and it switches the instrument to the power-on mode and causes the default measurement for the power-on mode to be active. This level of Restore System Defaults does not affect any system settings, but it does affect the state of all modes and does cause a mode switch unless the instrument was already in the power-on mode.

Confirmation is required to restore the factory default values. The confirmation dialog is:

.

Key Path

Example

System, Restore System Defaults

:SYST:DEF MOD

263

System Functions

System

Couplings An All Mode will cause the currently running measurement to be aborted, mode switch to the power-on mode and activate the default measurement for the power-on mode.. It gets the mode to a consistent state with all of the default couplings set.

Prior to A.02.00

Initial S/W Revision

All

This is the catastrophic function that does a comprehensive reset of ALL analyzer settings to their factory default values. It resets all of the system setting groups, causes a Restore Mode Defaults for all modes in the instrument, and switches back to the power-on mode. It does not affect the User Preset file or any user saved files.

Confirmation is required to restore the factory default values. The confirmation dialog is:

Key Path

Example

Couplings

Initial S/W Revision

System, Restore System Defaults

:SYST:DEF ALL

An All will cause the currently running measurement to be aborted and get all modes to a consistent state, so it is unnecessary to couple any settings.

Prior to A.02.00

Control Panel…

Opens the Windows Control Panel. The Control Panel is used to configure certain elements of Windows that are not configured through the hardkey/softkey System menus.

The Control Panel is a separate Windows application, so to return to the analyzer once you are in the

Control Panel, you may either:

Exit the Control Panel by clicking on the red X in the upper right hand corner, with a mouse

264

System Functions

System

Or use Alt-Tab: press and hold the Alt key and press and release the Tab key until the Analyzer logo is showing in the window in the center of the screen, as above, then release the Alt key.

Key Path

Notes

Initial S/W Revision

System

No remote command for this key.

Prior to A.02.00

Licensing…

Opens the license explorer.

For Help on this key, select Help in the menu bar at the top of the license explorer window.

Key Path

Notes

Initial S/W Revision

System

No equivalent remote command for this key.

Prior to A.02.00

Remote Command:

:SYSTem:LKEY <”OptionInfo”>, <”LicenseInfo”>

265

System Functions

System

Example:

Notes:

Initial S/W Revision:

Remote Command:

Example:

Notes:

Initial S/W Revision:

Remote Command:

SYST:LKEY

“N9073A–1FP”,”027253AD27F83CDA5673A9BA5F427FDA5E4F25AEB1

017638211AC9F60D9C639FE539735909C551DE0A91”

The <”OptionInfo”> contains the feature and the version. You must specify the feature but can omit the version. If you omit the version, the system regards it as the latest one, since the system knows which version is supported for each feature.

The <”LicenseInfo”> contains the signature, the expiration date, and serial number for transport if transportable. You must specify the signature, but you can omit the other information. If you omit the expiration date, the system regards it as permanent. If you omit the serial number, the system regards it as non-transportable. As a result, this supports backward compatibility.

Prior to A.02.00

:SYSTem:LKEY:DELete <”OptionInfo”>,<”LicenseInfo”>

SYST:LKEY:DEL

“N9073A–1FP”,”027253AD27F83CDA5673A9BA5F427FDA5E4F25AEB1

017638211AC9F60D9C639FE539735909C551DE0A91”

The <”OptionInfo”> contains the feature and the version. You must specify the feature but can omit the version. If you omit the version, the system regards it as the latest one, if more than one version is installed.

The <”LicenseInfo”> contains the signature, the expiration date, and whether or not be transportable. You must specify the signature, but you can omit the other information. If you omit the expiration date, the system regards it as permanent. If you omit the transportability, the system regards it as non-transportable. As a result, this supports backward compatibility.

Prior to A.02.00

:SYSTem:LKEY:LIST?

266

Notes:

Initial S/W Revision:

Remote Command:

Example:

Notes:

Initial S/W Revision:

Remote Command:

Notes:

Initial S/W Revision:

System Functions

System

Return Value:

An <arbitrary block data> of all the installed instrument licenses.

The format of each license is as follows.

<Feature>,<Version>,<Signature>,<Expiration Date>,<Serial Number for

Transport>

Return Value Example:

#3136

N9073A–1FP,1.000,B043920A51CA

N9060A–2FP,1.000,4D1D1164BE64

N9020A–508,1.000,389BC042F920

N9073A–1F1,1.000,5D71E9BA814C,13-aug–2005

<arbitrary block data> is:

#NMMM<data>

Where:

N is the number of digits that describes the number of MMM characters. For example if the data was 55 bytes, N would be 2.

MMM would be the ASCII representation of the number of bytes. In the previous example, N would be 55.

<data> ASCII contents of the data

Prior to A.02.00

:SYSTem:LKEY? <”OptionInfo”>

SYST:LKEY? “N9073A–1FP”

The <”OptionInfo”> contains the feature and the version. You must specify the feature but can omit the version. If you omit the version, the system regards it as the latest one.

Return Value:

<”LicenseInfo”> if the license is valid, null otherwise.

<”LicenseInfo”> contains the signature, the expiration date, and serial number if transportable.

Return Value Example:

“B043920A51CA”

Prior to A.02.00

:SYSTem:HID?

Return value is the host ID as a string

Prior to A.02.00

267

System Functions

System

Security

Accesses capabilities for operating the instrument in a security controlled environment.

Key Path

Initial S/W Revision

System

PXA.01.01

USB

The Windows operating system can be configured to disable write access to the USB ports for users who are in a secure environment where transferring data from the instrument is prohibited. This user interface is a convenient way for the customer to disable write access to USB.

Key Path

Mode

Scope

Remote Command

Example

Notes

Dependencies

Preset

State Saved

Range

Key Path

Initial S/W Revision

System, Security

All

Mode Global

:SYSTem:SECurity:USB:WPRotect[:ENABle] ON|OFF|0|1

:SYSTem:SECurity:USB:WPRotect[:ENABle]?

:SYST:SEC:USB:WPR ON ! Will set USB ports to Read-only

When the USB ports are in Read-only mode then no data can be stored to

USB, including the internal USB memory used for a back-up location for the calibration data.

This key is grayed-out unless the current user has administrator privileges.

This is unaffected by Preset or any Restore System Defaults. An Agilent

Recovery will set the USB to write protect OFF

No

Read-Write|Read only

3664

PXA.01.01

Read-Write

Selection for allowing full read-write access to the USB ports.

Key Path

Example

Initial S/W Revision

System, Security, USB

:SYST:SEC:USB:WPR OFF ! Will set USB ports to Read-Write

PXA.01.01

268

System Functions

System

Read only

Selection for disabling write access to the USB ports.

Key Path

Example

Initial S/W Revision

System, Security, USB

:SYST:SEC:USB:WPR ON ! Will set USB ports to Read only

PXA.01.01

Diagnostics

The Diagnostics key in the System menu gives you access to basic diagnostic capabilities of the instrument.

Key Path

Initial S/W Revision

System

Prior to A.02.00

Show Hardware Statistics

Provides a display of various hardware statistics. The statistics include the following:

Mechanical relay cycles

High and Low temperature extremes

Elapsed time that the instrument has been powered-on (odometer)

The display should appear listing the statistics, product number, serial number, and firmware revision.

Std Header

Product Number: N9020A

Serial Number: US46340924

Firmware Revision: A.01.01

Mechanical relays

Calibrator Switch Cycles:

AC/DC Switch Cycles:

2 dB #1 Mechanical Atten Cycles

2 dB #2 Mechanical Atten Cycles

6 dB Mechanical Atten Cycles

10 dB Mechanical Atten Cycles

20 dB Mechanical Atten Cycles

30 dB Mechanical Atten Cycles

Low Noise Path Switch

Preselector Bypass Cycles

1800

60

23489

23400

500000

1000000

2500

60000

4339

1000

In CXA this field is called “Fixed Atten”

Low Noise Path

Switch is only shown if Opt LNP hardware is installed

Preselector Bypass is only shown if Opt

MPB is installed

None of these fields appear in

CXA

Odometer

High operating temperature extreme:

Low operating temperature extreme

+37.2 degC

+18.1 degC

1600

The data will be updated only when the Show Hardware Statistics menu key is pressed, it will not be

269

System Functions

System

updated while the screen is displayed.

The tabular data should be directly printable.

Key Path

Mode

Notes

Initial S/W Revision

System, Diagnostics

All

The values displayed on the screen are only updated upon entry to the screen and not updated while the screen is being displayed.

Prior to A.02.00

SCPI for Show Hardware Statistics ( Remote Commands Only)

Each of the hardware statistic items can be queriedvia SCPI.

“Query the Mechanical Relay Cycle Count” on page 270

“Query the Operating Temperature Extremes” on page 270

“Query the Elapsed Time since 1st power on” on page 271

Query the Mechanical Relay Cycle Count

Returns the count of mechanical relay cycles.

Remote Command:

Example:

Notes:

Initial S/W Revision:

Modified at S/W Revision:

:SYSTem:MRELay:COUNt?

:SYST:MREL:COUN?

Query Only

The return value is a comma separated list of the individual counts for each mechanical relay.

The position of the relays in the list is:

“<Cal Signal>,<AC/DC>,<2dB #1 Atten>,<2dB #2 Atten>,<6dB

Atten>,<10dB Atten>,<20dB Atten>,<30dB Atten>,<Fixed Atten>,<Low

Noise Path Switch>,<Presel Bypass>”

Items in the list not pertaining to your particular hardware configuration will return as –999 for those items.

Prior to A.02.00

A.03.00, A.04.00

Query the Operating Temperature Extremes

Returns the low operating temperature extreme value. The value survives a power-cycle and is the temperature extreme encountered since the value was reset by the factory or service center.

Mode

Remote Command

Example

All

:SYSTem:TEMPerature:LEXTreme?

:SYST:TEMP:LEXT?

270

System Functions

System

Notes

State Saved

Initial S/W Revision

Value is in degrees Celsius at which the lowest operating temperature has been recorded since 1st power-up.

No

Prior to A.02.00

Mode

Remote Command

Example

Notes

State Saved

Initial S/W Revision

All

:SYSTem:TEMPerature:HEXTreme?

:SYST:TEMP:HEXT?

Value is in degrees Celsius at which the highest operating temperature has been recorded since 1st power-up.

No

Prior to A.02.00

Query the Elapsed Time since 1st power on

Returns the elapsed on-time in minutes since 1st power-on.

Remote Command:

Example:

Notes:

Initial S/W Revision:

:SYSTem:PON:ETIMe?

:SYST:PON:ETIM?

Query Only

Prior to A.02.00

Advanced

Accesses advanced diagnostic capabilities performed in the factory or under instructions from repair procedures. This menu key is only visible when the logged-in user is “saservice”. The first access to the

Advanced Diagnostic Menu after invoking the instrument application will require an authentication, which is to enter the Service Code. Subsequent accesses to the Advanced Diagnostic Menu are unimpeded. The Authentication dialog looks like:

“OK” is the default key thus the Enter key is used to complete the entry. If invalid Service Code is entered authentication is not granted and you are provided the following dialog:

271

System Functions

System

Key Path

Notes

Initial S/W Revision

System, Diagnostics

Password is required to access this menu.

Prior to A.02.00

Service

Accesses capabilities performed in the factory or under instructions from repair procedures. This menu key is only visible when the logged-in user is “advanceduser” or “saservice”. The first access to the

Service Menu after invoking the instrument application will require an authentication Service Code.

Key Path

Initial S/W Revision

System

Prior to A.02.00

System Remote Commands (Remote Commands Only)

The commands in this section have no front panel key equivalent

Initial S/W Revision: Prior to A.02.00

List installed Options (Remote Command Only)

Lists the installed options that pertain to the instrument (signal analyzer). .

Mode

Remote Command

Example

Notes

State Saved

Initial S/W Revision

All

:SYSTem:OPTions?

:SYST:OPT?

The return string is a comma separated list of the installed options. For example:

“503,P03,PFR”

:SYSTem:OPTions? and *OPT? are the same.

No

Prior to A.02.00

272

System Functions

System

Lock the Front-panel keys (Remote Command Only)

Disables the instrument keyboard to prevent local input when the instrument is controlled remotely.

Annunciation showing a “K” for ‘Klock” (keyboard lock) alerts the local user that the keyboard is locked. Klock is similar to the GPIB Local Lockout function; namely that no front-panel keys are active with the exception of the Power Standby key. (The instrument is allowed to be turned-off if Klock is

ON.) The Klock command is used in remote control situations where Local Lockout cannot be used.

Although primary intent of Klock is to lock-out the front panel, it will lock-out externally connected keyboards through USB. Klock has no effect on externally connected pointing devices (mice).

The front panel ‘Local’ key (Cancel/Esc) has no effect if Klock is ON.

Mode

Remote Command

Example

Notes

Preset

State Saved

Initial S/W Revision

All

:SYSTem:KLOCk OFF|ON|0|1

:SYSTem:KLOCk?

:SYST:KLOC ON

Keyboard lock remains in effect until turned-off or the instrument is power-cycled

Initialized to OFF at startup, unaffected by Preset

No

Prior to A.02.00

List SCPI Commands (Remote Command Only)

Outputs a list of the valid SCPI commands for the currently selected Mode.

Remote Command:

Example:

Notes:

Initial S/W Revision:

:SYSTem:HELP:HEADers?

:SYST:HELP:HEAD?

The output is an IEEE Block format with each command separated with the

New-Line character (hex 0x0A)

Prior to A.02.00

SCPI Version Query (Remote Command Only)

Returns the SCPI version number with which the instrument complies. The SCPI industry standard changes regularly. This command indicates the version used when the instrument SCPI commands were defined.

Remote Command:

Example:

Initial S/W Revision:

:SYSTem:VERSion?

:SYST:VERS?

Prior to A.02.00

273

System Functions

System

Date (Remote Command Only)

The recommended access to the Date, Time, and Time zone of the instrument is through the Windows native control (Control Panel or accessing the Task Bar). You may also access this information remotely, as shown in this command and Time (below).

Sets or queries the date in the instrument.

Mode

Remote Command

Example

Notes

Initial S/W Revision

All

:SYSTem:DATE “<year>,<month>,<day>”

:SYSTem:DATE?

:SYST:DATE “2006,05,26”

<year> is the four digit representation of year. (for example, 2006)

<month> is the two digit representation of year. (for example. 01 to 12)

<day> is the two digit representation of day. (for example, 01 to 28, 29, 30, or

31) depending on the month and year

Prior to A.02.00

Time (Remote Command Only)

Sets or queries the time in the instrument.

Mode

Remote Command

Example

Notes

Initial S/W Revision

All

:SYSTem:TIME “<hour>,<minute>,<second>”

:SYSTem:TIME?

:SYST:TIME “13,05,26”

<hour> is the two digit representation of the hour in 24 hour format

<minute> is the two digit representation of minute

<second> is the two digit representation of second

Prior to A.02.00

274

System Functions

User Preset

User Preset

Accesses a menu that gives you the following three choices:

User Preset – recalls a state previously saved

using the

Save User Preset

function.

User Preset All Modes – presets

all of the modes in the analyzer

Save User Preset –

saves the current state for the current mode

Key Path

Initial S/W Revision

Front-panel key

Prior to A.02.00

User Preset

User Preset

sets the state of the currently active mode back to the state that was previously saved for this mode using the

Save User Preset

menu key or the SCPI command, SYST:PRES:USER:SAV. It not only recalls theMode Preset settings, but it also recalls all of the mode persistent settings, and the

Input/Outpit system setting that existed at the time

Save User Preset

was executed.

If a

Save User Preset

has not been done at any time,

User Preset

recalls the default user preset file for the currently active mode. The default user preset files are created if, at power-on, a mode detects there is no user preset file. There will never be a scenario when there is no user preset file to restore. For each mode, the default user preset state is the same state that would be saved if a

Save User Preset

is performed in each mode right after doing a Restore Mode Default and after a Restore Input/Output

Defaults.

The User Preset function does the following:

• Aborts the currently running measurement.

• Sets the mode State to the values defined by

Save User Preset

.

• Makes the saved measurement for the currently running mode the active measurement.

• Brings up the saved menu for the power-on mode.

• Clears the input and output buffers.

• Sets the Status Byte to 0.

Key Path

Remote Command

Example

User Preset

:SYSTem:PRESet:USER

:SYST:PRES:USER:SAVE

:SYST:PRES:USER

275

System Functions

User Preset

Notes

Couplings

Initial S/W Revision

:SYST:PRES:USER:SAVE is used to save the current state as the user preset state.

Clears all pending OPC bits. The Status Byte is set to 0.

Pressing the User Preset front-panel key while already in the User Preset menu will cause the User Preset to get executed

A user preset will cause the currently running measurement to be aborted and cause the saved measurement to be active. Recalling a User Preset file has the same issues that recalling a Save State file has. Some settings may need to be limited and therefore re-coupled, since the capabilities of the mode may have changes when the User Preset file was last saved.

Prior to A.02.00

User Preset All Modes

Recalls all of the User Preset files for each mode, switches to the power-on mode, and activates the saved measurement from the power-on mode User Preset file.

NOTE

When the instrument is secured, all of the user preset files are converted back to their default user preset files.

The User Preset function does the following:

• Aborts the currently running measurement.

• Switches the Mode to the power-on mode.

• Restores the User Preset files for each mode.

• Makes the saved measurement for the power-on mode the active measurement.

• Brings up the saved menu for the power-on mode.

• Clears the input and output buffers.

• Sets the Status Byte to 0.

Key Path

Remote Command

Example

Notes

User Preset

:SYSTem:PRESet:USER:ALL

:SYST:PRES:USER:SAVE

:SYST:PRES:USER:ALL

Clears all pending OPC bits. The Status Byte is set to 0.

:SYST:PRES:USER:SAVE is used to save the current state as the user preset state.

276

System Functions

User Preset

Couplings

Initial S/W Revision

A user preset will cause the currently running measurement to be aborted, cause a mode switch to the power-on mode, and cause the saved measurement to be active in the power-on mode. Recalling a User Preset file has the same issues that recalling a Save State file has. Some settings may need to be limited and therefore re-coupled, since the capabilities of the mode may have changes when the User Preset file was last saved.

Prior to A.02.00

Save User Preset

Saves the currently active mode and its State. You can recall this User Preset file by pressing the User

Preset menu key or sending the SYST:PRES:USER remote command. This same state is also saved by the Save State function.

Key Path

Remote Command

Example

Notes

Initial S/W Revision

User Preset

:SYSTem:PRESet:USER:SAVE

:SYST:PRES:USER:SAVE

:SYST:PRES:SAVE creates the same file as if the user requested a *SAV or a

MMEM: STOR:STAT, except User Preset Save does not allow the user to specify the filename or the location of the file.

Prior to A.02.00

277

System Functions

User Preset

278

6

Programming the Analyzer

This chapter provides introductory information about the programming documentation included with your product.

279

Programming the Analyzer

What Programming Information is Available?

What Programming Information is Available?

The X-Series Documentation can be accessed through the Additional Documentation page in the instrument Help system and is included on the Documentation CD shipped with the instrument. It can also be found in the instrument at: C:\ProgramsFiles\Agilent\SignalAnalysis\

Infrastructure\Help\otherdocs

, or online at: http://www.agilent.com/find/mxa_manuals.

The following resources are available to help you create programs for automating your X-Series measurements:

Resource Description

X-Series

Programmer’s Guide

Provides general SCPI programming information on the following topics:

• Programming the X-Series Applications

• Programming fundamentals

• Programming examples

Note that SCPI command descriptions for measurement applications are NOT in this book, but are in the User’s and Programmer’s Reference.

User’s and

Programmer’s

Reference manuals

Describes all front-panel keys and softkeys, including SCPI commands for a measurement application. Note that:

Embedded Help in your instrument

• Each measurement application has its own User’s and Programmer’s Reference.

• The content in this manual is duplicated in the analyzer’s Help (the Help that you see for a key is identical to what you see in this manual).

Describes all front-panel keys and softkeys, including SCPI commands, for a measurement application.

Note that the content that you see in Help when you press a key is identical to what you see in the User’s and Programmer’s Reference.

X-Series Getting

Started Guide

Provides valuable sections related to programming including:

• Licensing New Measurement Application Software - After Initial Purchase

• Configuring instrument LAN Hostname, IP Address, and Gateway Address

• Using the Windows XP Remote Desktop to connect to the instrument remotely

• Using the Embedded Web Server Telnet connection to communicate SCPI

This printed document is shipped with the instrument.

Agilent Application

Notes

Printable PDF versions of pertinent application notes.

Agilent VISA User’s

Guide

Describes the Agilent Virtual Instrument Software Architecture (VISA) library and shows how to use it to develop I/O applications and instrument drivers on Windows PCs.

280 Chapter 6

Programming the Analyzer

IEEE Common GPIB Commands

IEEE Common GPIB Commands

Numeric values for bit patterns can be entered using decimal or hexi-decimal representations. (that is,. 0 to 32767 is equivalent to #H0 to #H7FFF).

Calibration Query

*CAL? Performs a full alignment and returns a number indicating the success of the alignment. A zero is returned if the alignment is successful. A one is returned if any part of the alignment fails. The equivalent

SCPI command is CALibrate[:ALL]?

See “Alignments” on page 193 for details of *CAL?.

Clear Status

Clears the status byte register. It does this by emptying the error queue and clearing all bits in all of the event registers. The status byte register summarizes the states of the other registers. It is also responsible for generating service requests.

Key Path

Remote Command

Example

Notes

Status Bits/OPC dependencies

Initial S/W Revision

No equivalent key. Related key System, Show Errors, Clear Error Queue

*CLS

*CLS Clears the error queue and the Status Byte Register.

For related commands, see the SYSTem:ERRor[:NEXT]? command. See also the STATus:PRESet command and all commands in the STATus subsystem.

Resets all bits in all event registers to 0, which resets all the status byte register bits to 0 also.

Prior to A.02.00

Standard Event Status Enable

Selects the desired bits from the standard event status enable register. This register monitors I/O errors and synchronization conditions such as operation complete, request control, query error, device dependent error, status execution error, command error, and power on. The selected bits are OR’d to become a summary bit (bit 5) in the byte register which can be queried.

The query returns the state of the standard event status enable register.

Key Path

Remote Command

No equivalent key. Related key System, Show Errors, Clear Error Queue

*ESE <integer>

*ESE?

Chapter 6 281

Programming the Analyzer

IEEE Common GPIB Commands

Example

Notes

Preset

State Saved

Min

Max

Status Bits/OPC dependencies

Initial S/W Revision

*ESE 36 Enables the Standard Event Status Register to monitor query and command errors (bits 2 and 5).

*ESE? Returns a 36 indicating that the query and command status bits are enabled.

For related commands, see the STATus subsystem and

SYSTem:ERRor[:NEXT]? commands.

255

Not saved in state.

0

255

Event Enable Register of the Standard Event Status Register.

Prior to A.02.00

Standard Event Status Register Query

Queries and clears the standard event status event register. (This is a destructive read.) The value returned is a hexadecimal number that reflects the current state (0/1) of all the bits in the register.

Remote Command:

Example:

Notes:

Preset:

Min:

Max:

Status Bits/OPC dependencies:

Initial S/W Revision:

*ESR?

*ESR? Returns a 1 if there is either a query or command error, otherwise it returns a zero.

For related commands, see the STATus subsystem commands.

0

0

255

Standard Event Status Register (bits 0 – 7).

Prior to A.02.00

Identification Query

Returns a string of instrument identification information. The string will contain the model number, serial number, and firmware revision.

The response is organized into four fields separated by commas. The field definitions are as follows:

• Manufacturer

• Model

• Serial number

282 Chapter 6

Programming the Analyzer

IEEE Common GPIB Commands

• Firmware version

Key Path

Remote Command

Example

Initial S/W Revision

No equivalent key. See related key System, Show System.

*IDN?

*IDN? Returns instrument identification information, such as:

Agilent Technologies,N9020A,US01020004,A.01.02

Prior to A.02.00

Instrument Model Number

ID? - Returns a string of the instrument identification. The string will contain the model number.

When in Remote Language compatibility mode the query will return the model number of the emulated instrument, when in any other mode the returned model number will be that of the actual hardware.

Operation Complete

The *OPC command sets bit 0 in the standard event status register (SER) to “1” when pending operations have finished, that is when all overlapped commands are complete. It does not hold off subsequent operations. You can determine when the overlapped commands have completed either by polling the OPC bit in SER, or by setting up the status system such that a service request (SRQ) is asserted when the OPC bit is set.

The *OPC? query returns a “1” after all the current overlapped commands are complete. So it holds off subsequent commands until the "1” is returned, then the program continues. This query can be used to synchronize events of other instruments on the external bus.

Remote Command:

Example:

Status Bits/OPC dependencies:

Initial S/W Revision:

*OPC

*OPC?

INIT:CONT 0 Selects single sweeping.

INIT:IMM Initiates a sweep.

*OPC? Holds off any further commands until the sweep is complete.

Not global to all remote ports or front panel. *OPC only considers operation that was initiated on the same port as the *OPC command was issued from.

*OPC is an overlapped command, but *OPC? is sequential.

Prior to A.02.00

Query Instrument Options

Returns a string of all the installed instrument options. It is a comma separated list with quotes, such as:

“503,P03,PFR”.

To be IEEE compliant, this command should return an arbitrary ascii variable that would not begin and end with quotes. But the quotes are needed to be backward compatible with previous SA products and

Chapter 6 283

Programming the Analyzer

IEEE Common GPIB Commands

software. So, the actual implementation will use arbitrary ascii. But quotes will be sent as the first and last ascii characters that are sent with the comma-separated option list.

Remote Command:

Initial S/W Revision:

*OPT?

Prior to A.02.00

Recall Instrument State

This command recalls the instrument state from the specified instrument memory register.

• If the state being loaded has a newer firmware revision than the revision of the instrument, no state is recalled and an error is reported.

• If the state being loaded has an equal firmware revision than the revision of the instrument, the state will be loaded.

• If the state being loaded has an older firmware revision than the revision of the instrument, the instrument will only load the parts of the state that apply to the older revision.

Remote Command:

Example:

Notes:

Min:

Max:

Status Bits/OPC dependencies:

Initial S/W Revision:

*RCL <register #>

*RCL 7 Recalls the instrument state that is currently stored in register 7.

Registers 0 through 6 are accessible from the front panel in menu keys for

Recall Registers.

0

127

The command is sequential.

Prior to A.02.00

Save Instrument State

This command saves the current instrument state and mode to the specified instrument memory register.

Remote Command:

Example:

Notes:

Min:

Max:

Status Bits/OPC dependencies:

Initial S/W Revision:

*SAV <register #>

*SAV 9 Saves the instrument state in register 9.

Registers 0 through 6 are accessible from the front panel in menu keys for

Save Registers.

0

127

The command is sequential.

Prior to A.02.00

284 Chapter 6

Programming the Analyzer

IEEE Common GPIB Commands

Service Request Enable

This command enables the desired bits of the service request enable register.

The query returns the value of the register, indicating which bits are currently enabled.

Remote Command:

Example:

Notes:

Preset:

Min:

Max:

Status Bits/OPC dependencies:

Initial S/W Revision:

*SRE <integer>

*SRE?

*SRE 22 Enables bits 1, 2, and 4 in the service request enable register.

For related commands, see the STATus subsystem and

SYSTem:ERRor[:NEXT]? commands.

0

0

255

Service Request Enable Register (all bits, 0 – 7).

Prior to A.02.00

Status Byte Query

Returns the value of the status byte register without erasing its contents.

Remote Command:

Example:

Notes:

Status Bits/OPC dependencies:

Initial S/W Revision:

*STB?

*STB? Returns a decimal value for the bits in the status byte register.

For example, if a 16 is returned, it indicates that bit 5 is set and one of the conditions monitored in the standard event status register is set.

See related command *CLS.

Status Byte Register (all bits, 0 – 7).

Prior to A.02.00

Trigger

This command triggers the instrument. Use the :TRIGger[:SEQuence]:SOURce command to select the trigger source.

Key Path

Remote Command

Example

Notes

Initial S/W Revision

No equivalent key. See related keys Single and Restart.

*TRG

*TRG Triggers the instrument to take a sweep or start a measurement, depending on the current instrument settings.

See related command :INITiate:IMMediate.

Prior to A.02.00

Chapter 6 285

Programming the Analyzer

IEEE Common GPIB Commands

Self Test Query

This query performs the internal self-test routines and returns a number indicating the success of the testing. A zero is returned if the test is successful, 1 if it fails.

Remote Command:

Example:

Initial S/W Revision:

*TST?

*TST? Runs the self-test routines and returns 0=passed, 1=some part failed.

Prior to A.02.00

Wait-to-Continue

This command causes the instrument to wait until all overlapped commands are completed before executing any additional commands. There is no query form for the command.

Remote Command:

Example:

Status Bits/OPC dependencies:

Initial S/W Revision:

*WAI

INIT:CONT OFF; INIT;*WAI Sets the instrument to single sweep. Starts a sweep and waits for its completion.

Not global to all remote ports or front panel. *OPC only considers operation that was initiated on the same port as the *OPC command was issued from.

Prior to A.02.00

286 Chapter 6

7

Vector Analysis

The Vector Analysis measurement is accessed from the Meas hardkey.The Vector Analysis measurement bases its results on a set of periodic time samples of a channel. The channel is defined by a combined bank of hardware and DSP filters whose overall frequency response has a flat top and steep rolloff at the band edges. The time record is operated upon by a number of mathematical functions, including the FFT to produce spectrum results and statistical functions, including complementary cumulative distribution function (CCDF). Any of these results may be displayed in a flexible layout, with the Y data formatted in a variety of ways, and results scaled as desired. Many of these analysis results are also available in optional VSA demodulation measurements. For measurement results and

views, see “View/Display” on page 311 .

This topic contains the following sections:

“Remote Command Results for Vector Analysis Measurement” on page 287

“Front Panel Results” on page 287

Remote Command Results for Vector Analysis Measurement

The Vector Analysis measurement is invoked remotely by the following:

:CONFigure:VECTor

:CONFigure:VECTor:NDEFault

Remote results may be obtained using the following:

:FETCh:VECTor[n]?

Only table results may be obtained using FETCh. The tables available for the Vector Analysis measurement are ACP and OccBW tables, which are available to any VSA measurement.

:INITiate:VECTor

:READ:VECTor[n]?

NOTE

The MEASure? command is not supported by the Vector Analysis measurement.

For more information and remote commands, see “Remote SCPI Commands and Data Queries” on page 482

in the section, “Common Measurement Functions 2”.

Also see

“Data” on page 430

in the section “Common Measurement Functions 2” for more measurement

SCPI commands.

Front Panel Results

Vector Analysis results may be displayed in any trace, and the traces viewed in a variety of layouts that show 1, 2, 3, or 4 traces at a time. Each trace may be scaled as desired regardless of measurement settings, or auto-scaled to reflect measurement settings. Data may be formatted in a variety of ways. (For example, you may view the log magnitude of complex data, the real or imaginary part, etc.) You may use View Presets to view frequently used results, or to provide a familiar starting point from which you

287

Vector Analysis may customize your own view.

Key Path

Mode

Meas

VSA

288

Vector Analysis

AMPTD Y Scale

AMPTD Y Scale

See “AMPTD Y Scale (Amplitude)” on page 361 in the section "Common Measurement Functions 2"

for for a description of this function.

289

Vector Analysis

Auto Couple

Auto Couple

The Auto Couple key forces all Auto/Man functions into Auto. These include the following functions.

• Frequency Step

• X Scale

• Y Axis Unit Preference

• Frequency Points

NOTE

This function does not invoke the Auto tune function, nor does it cause any Y autoscaling or Input Auto ranging.

For more information, see

“Auto Couple” on page 33

in the section "Common Measurement Functions".

Key Path

Mode

Front Panel

VSA

290

Vector Analysis

BW

BW

See “BW (Bandwidth)” on page 368 in the section "Common Measurement Functions 2" for a

description of this function.

291

Vector Analysis

Cont (Continuous)

Cont (Continuous)

See

“Cont (Continuous Measurement/Sweep)” on page 35

in the section "Common Measurement

Functions 1" for a description of this function.

292

Vector Analysis

FREQ Channel

FREQ Channel

See “FREQ Channel” on page 371

in the section "Common Measurement Functions 2" for a description of this function.

293

Vector Analysis

Input/Output

Input/Output

See

“Input/Output” on page 43

in the section "Common Measurement Functions " for a description of this function.

294

Vector Analysis

Marker

Marker

See “Marker” on page 374

in the section "Common Measurement Functions 2" for a description of this function.

295

Vector Analysis

Marker Function

Marker Function

See

“Marker Function” on page 392

in the section "Common Measurement Functions 2" for a description of this function.

296

Vector Analysis

Marker To

Marker To

See “Marker -> (Marker To)” on page 390 in the section "Common Measurement Functions 2" for a

description of this function.

297

Vector Analysis

Meas (Measure)

Meas (Measure)

See

“Meas” on page 155 in the section "Common Measurement Functions " for a description of this

function.

298

Vector Analysis

Meas Setup

Meas Setup

See “Meas Setup” on page 400

in the section "Common Measurement Functions 2" for a description of this function.

Avg Number

See “Avg Number” on page 400 in the section "Common Measurement Functions 2" for a description of

this function.

Average Mode

See “Average Mode” on page 401 in the section "Common Measurement Functions 2" for a description

of this function.

Average Setup

See “Average Setup” on page 402 in the section "Common Measurement Functions 2" for a description

of this function.

Meas Preset

Returns parameters for this measurement to those set by the factory.

Key Path

Mode

Remote Command

Example

Meas Setup

VXA

:CONFigure:VECTor

CONF:VECT

299

Vector Analysis

Mode

Mode

See

“Mode” on page 213 in the section "Common Measurement Functions" for a description of this

function.

300

Vector Analysis

Mode Setup

Mode Setup

See “Mode Setup” on page 229 in the section "Common Measurement Functions" for a description of

this function.

301

Vector Analysis

Peak Search

Peak Search

See

“Peak Search” on page 406

in the section "Common Measurement Functions 2" for a description of this function.

302

Vector Analysis

Single

Single

See “Single (Single Measurement/Sweep)” on page 263

in the section "Common Measurement

Functions" for a description of this function.

303

Vector Analysis

Source

Source

See

“Source” on page 263 in the section "Common Measurement Functions" for a description of this

function.

304

Vector Analysis

SPAN X Scale

SPAN X Scale

See “SPAN X Scale” on page 416

in the section "Common Measurement Functions 2" for a description of this function.

305

Vector Analysis

Sweep/Control

Sweep/Control

See

“Sweep/Control” on page 424

in the section "Common Measurement Functions 2" for a description of this function.

306

Vector Analysis

Trace/Detector

Trace/Detector

This section details the trace results accessible via the data key and also via SCPI, many of which are available in other measurements as well.

See “Trace/Detector” on page 428 in the section "Common Measurement Functions 2" for a description

of this function.

Select Trace

See “Select Trace” on page 428 in the section "Common Measurement Functions 2" for a description of

this function.

Data

This key provides a menu of trace data choices for the selected trace. For the SCPI command, and other details, see Trace/Detector,

“Data” on page 430

.

The following trace data results are available:

Trace data soft key name

SCPI string form

Spectrum "Spectrum1"

Description

Inst Spectrum

Main Time

Inst Main Time

Gate Time

Raw Main Time

"Inst Spec1"

"Main Time1"

"Inst Main Time1"

"Gate Time1"

"Raw Main Time1"

"PSD1"

Averaged result of successive Inst Spectrum results (If RMS or Max averaging is on). Otherwise, FFT of current windowed Main (or Gate)

Time

Instantaneous Spectrum is the FFT of the current windowed Main (or

Gate) Time. It is instantaneous in the sense that it isn't rms averaged, but it may included time-averaged data.

Same as Inst Main Time unless Time averaging is on, in which case it is the averaged result of successive Inst Main Time results.

If Time Gating is on, Main Time is the source to which the Gate is applied.

Instantaneous Main Time is the current corrected, resampled, time record.

Gate Time replaces Main Time as input to results if gating is on.

Time record as it comes from the hardware, before software resampling or corrections

Power spectrum divided by ResBW Power Spectral

Density (PSD)

CCDF "CCDF1"

CDF "CDF1"

Complementary Cumulative Distribution Function of all time date since last measurement restart

Cumulative Distribution Function of all time date since last measurement restart

307

Vector Analysis

Trace/Detector

Trace data soft key name

SCPI string form

PDF

Auto Correlation

OBW Summary

Trace 1

ACP Summary

Trace 1

No Data

Description

"PDF1" Probability Distribution Function of all time date since last measurement restart

Autocorrelation function of the current Main (or Gate) Time result "Auto Correl1"

"OBW Summary Trc1" Table of Occupied Bandwidth results if OccBW is enabled on Trace 1 and Trace 1 has Spectrum or PSD data. Similar summaries are available for all traces.

"ACP Summary Trc1" Table of Adjacent Channel Power results if ACP is enabled on Trace

1 and Trace 1 has Spectrum or PSD data. Similar summaries are available for all traces.

"No Data" An empty trace

Key Path

Mode

Trace/Detector

VSA

Format

This key provides a menu of Trace Formats from which to choose for the selected trace.

See Trace/Detector, “Format” on page 443 in the section "Common Measurement Functions 2" for more

information.

Digital Demod Trace Setup

This key provides a menu that allows you to set Digital Demod parameters for the selected trace.

See Trace/Detector,

“Digital Demod Trace Setup” on page 444

in the section "Common Measurement

Functions 2" for more information.

Copy to Data Register

This key provides a menu of Data Registers numbered from 1 to 6 into which to copy the selected trace.

See Trace/Detector,

“Copy to Data Register” on page 450 in the section "Common Measurement

Functions 2" for more information.

Phase/Delay Properties

This key provides a menu that allows you to set Phase and Delay properties for the selected trace.

See Trace/Detector,

“Phase/Delay Properties” on page 451

in the section "Common Measurement

Functions 2" for more information.

308

Vector Analysis

Trace/Detector

ACP Setup

This key provides a menu that allows you to set ACP setup parameters for the selected trace.

See Trace/Detector, “ACP Setup” on page 453

in the section "Common Measurement Functions 2" for more information.

OBW Setup

This key provides a menu that allows you to set OBW setup parameters for the selected trace.

See Trace/Detector, “OBW Setup (Occupied Bandwidth)” on page 463

in the section "Common

Measurement Functions 2" for more information.

Trace Indicator Info

Pressing this key invokes the display of the currently selected trace indicator in the MSG bar at the bottom of the screen.

See Trace/Detector,

“Trace Indicator Info” on page 469

in the section "Common Measurement Functions

2" for more information.

309

Vector Analysis

Trigger

Trigger

Triggering is used to determine when a measurement should start taking data. There are several available trigger sources. For each trigger source, there are associated setup parameters. Typically, a trigger event is generated when a signal (or a characteristic of the signal) crosses a defined trigger level (or threshold) on a rising or falling slope. The measurement begins at a specified time delay from the trigger point. The delay may be negative, allowing pretrigger data to be taken. Each trigger source has associated its own trigger level, slope, and delay settings.

See

“Trigger” on page 470

in the section "Common Measurement Functions 2" for a description of this function.

310

Vector Analysis

View/Display

View/Display

The View/Display key provides access to a menu that enable you to select display parameters for the current measurement.

View Presets affect the trace layout, trace data assignment, scaling and formatting but do not affect hardware measurement setup.

Remote Command

Example

Key Path

Mode

:DISPlay:VECTor:VIEW:PRESet SPECtrum|STATistics

:DISP:VECT:VIEW:PRES SPEC

Front Panel

VSA

Display

See “Display” on page 349 in the section "Common Measurement Functions" for a description of this

function.

Layout

See “Layout” on page 479

in the section "Common Measurement Functions 2" for a description of this function.

Preset View: Spectrum/Time

This preset uses the Stack 2 layout style (see View/Display, Layout) with Spectrum in trace 1 and Main

Time in trace 2

Key Path

Mode

View/Display

VSA

311

Vector Analysis

View/Display

Preset View: Statistics

This preset uses the Stack 2 layout style (see View/Display, Layout) with the CCDF in trace 1 and Main

Time in trace 2.

Key Path

Mode

View/Display

VSA

312

Vector Analysis

View/Display

313

Vector Analysis

View/Display

314

8

Analog Demod

The Analog Demod measurement is accessed from the Meas hardkey. The Analog Demod measurement enhances the capabilities offered by Vector Analysis by adding a demodulation function. AM, FM, and

PM signals may be demodulated;, and the output of the demodulator may be further analyzed in the same ways as the input signal. For example, you may look simultaneously at the spectra and time records of the input signal and the demodulated signal. Also available are time gating, autocorrelation, and statistical functions such as CCDF.

This topic contains the following sections:

“Remote Command Results for Analog Demod Measurement” on page 315

“Front Panel Results” on page 315

Remote Command Results for Analog Demod Measurement

The Analog Demod measurement is invoked remotely by the following:

:CONFigure:ADEMod

:CONFigure:ADEMod:NDEFault

:INITiate:ADEMod

Remote results may be obtained using commands from the CALCulate:ADEMod:DATA<n> tree. In addition, the FETCh and READ commands below may be used to obtain ACP and OBW table information when those functions are turned on.

:FETCh:ADEMod[n]?

:READ:ADEMod[n]?

The MEASure? command is not supported by the Analog Demod measurement.

For more information and remote commands, see “Remote SCPI Commands and Data Queries” on page 482

in the section, “Common Measurement Functions 2”.

Also see Trace/Detector,

“Data” on page 430

in the section “Common Measurement Functions 2” for more measurement SCPI commands.

Front Panel Results

Analog Demod results may be displayed in any trace, and the traces viewed in a variety of layouts that show 1, 2, 3, or 4 traces at a time. Each trace may be scaled as desired regardless of measurement settings, or auto-scaled to reflect measurement settings. Data may be formatted in a variety of ways. (For example, you may view the log magnitude of complex data, the real or imaginary part, etc.) You may use

View Presets to view frequently used results, or to provide a familiar starting point from which you may customize your own view.

Key Path

Mode

Meas

VSA

315

Analog Demod

AMPTD Y Scale

AMPTD Y Scale

See

“AMPTD Y Scale (Amplitude)” on page 361

in the section "Common Measurement Functions 2" for a description of this function.

316

Analog Demod

Auto Couple

Auto Couple

See “Auto Couple” on page 33

in the section "Common Measurement Functions" for a description of this function.

317

Analog Demod

BW

BW

See

“BW (Bandwidth)” on page 368

in the section "Common Measurement Functions 2" for a description of this function.

318

Analog Demod

Cont (Continuous)

Cont (Continuous)

See “Cont (Continuous Measurement/Sweep)” on page 35 in the section "Common Measurement

Functions 1" for a description of this function.

319

Analog Demod

FREQ Channel

FREQ Channel

See

“FREQ Channel” on page 371 in the section "Common Measurement Functions 2" for a description

of this function.

320

Analog Demod

Input/Output

Input/Output

See “Input/Output” on page 43 in the section "Common Measurement Functions " for a description of

this function.

321

Analog Demod

Marker

Marker

See

“Marker” on page 374 in the section "Common Measurement Functions 2" for a description of this

function.

322

Analog Demod

Marker Function

Marker Function

See “Marker Function” on page 392 in the section "Common Measurement Functions 1" for a

description of this function.

323

Analog Demod

Marker To

Marker To

See

“Marker -> (Marker To)” on page 390

in the section "Common Measurement Functions 1" for a description of this function.

324

Analog Demod

Meas (Measure)

Meas (Measure)

See “Meas” on page 155

in the section "Common Measurement Functions " for a description of this function.

325

Analog Demod

Meas Setup

Meas Setup

The common Meas Setup functions are documented in “Meas Setup” on page 400 in the section

"Common Measurement Functions 2".

Avg Number

See

“Avg Number” on page 400

in the section "Common Measurement Functions 2" for a description of this function.

Average Mode

See

“Average Mode” on page 401

in the section "Common Measurement Functions 2" for a description of this function.

Average Setup

See

“Average Setup” on page 402

in the section "Common Measurement Functions 2" for a description of this function.

Demod Setup

Allows you to set Demodulation parameters for the current measurement.

For more information see: Measurement Functions, Meas Setup, Demod Setup

Key Path

Mode

Meas Setup

VSA

Demod Type

Demod Type lets you select the type of analog demodulation to be applied to your signal. You can select

AM, PM, or FM demodulation. In addition, you can enable Auto Carrier Phase and/or Auto Carrier

Frequency. Auto carrier controls how the analyzer determines your carrier frequency.

Remote Command

Example

Key Path

Mode

Preset

State Saved

[:SENSe]:ADEMod:MODulation AM|FM|PM

[:SENSe]:ADEMod:MODulation?

ADEM:MOD AM

ADEM:MOD?

Meas Setup, Demod Setup

VSA

AM

Saved in instrument state.

326

Analog Demod

Meas Setup

Range AM | FM | PM

AM Units

This allows you to select whether the display units for AM demodulation are in normalized units ("am") or percent ("%").

Remote Command

Restriction and Notes

Example

Key Path

Mode

Preset

State Saved

Range

[:SENSe]:ADEMod:AM:UNIT AM|PCT

[:SENSe]:ADEMod:AM:UNIT?

Greyed out if Demod Type is FM or PM

ADEM:AM:UNIT AM

ADEM:AM:UNIT?

Meas Setup, Demod Setup

VSA

AM

Saved in instrument state.

am | %

Auto Carrier Freq

This allows you to turn on or off automatic carrier frequency estimation for FM or PM demodulation.

When Auto Carrier Freq is off, the analyzer uses the Center Frequency setting as the carrier. If the actual carrier frequency is different from the center frequency, a ramp is visible in the phase results. Turning on

Auto Carrier Freq causes the analyzer to estimate the actual carrier frequency. Propper setting of the carrier frequency is especially important in PM demodulation.

When the VSA is in analog demodulation mode, you can select one of two different types of auto carrier setting. The types of auto carrier setting available depends on the type of demodulation, as shown in the following table.

For this demodulation:

AM

FM

PM

You can use this auto carrier:

None

Auto Carrier Frequency

Auto Carrier Frequency and Auto Carrier Phase

Note that you cannot select auto carrier frequency with AM demodulation. AM demodulation does not require carrier frequency estimation because the AM calculations are based on the carrier envelope. The carrier amplitude estimate is based on data within a single time record and is updated on a record-by-record basis.

Auto Carrier Frequency

Accurate angle demodulation (FM or PM) depends on precisely identifying the carrier frequency. Errors

327

Analog Demod

Meas Setup

result in phase ramping. The arc tangent of the complex time record is the basis of both PM and FM demodulation. Hence, correcting for the phase ramp is the goal of auto carrier frequency.

Without auto carrier frequency, the analyzer uses its center frequency to determine the carrier frequency of your signal. When auto carrier frequency is selected, the analyzer uses an algorithm to estimate the carrier frequency. If you can lock the analyzer to an external reference which is coherent with your carrier, no carrier frequency estimation is needed, and you do not need to select auto carrier frequency.

When auto carrier frequency is selected, the carrier frequency estimate is calculated independently for each time record, and is used in the demodulation calculation to take out FM offsets, or PM phase ramps due to error between your carrier frequency and the analyzer's LO (center frequency).

Hint

If auto carrier frequency is selected, you can select demod carrier to display the estimated carrier frequency for FM demodulation. Cross channel results show the estimates for both channels. For PM demodulation, you must select both Auto Carrier Frequency and Auto Carrier Phase to display the estimated carrier frequency.

Auto Carrier Frequency and Averaging

The operation of auto carrier frequency is modified if averaging is turned on. For FM measurements, the carrier frequency estimate from the current time record is fed into an exponential average of estimates from prior time records. The resulting, averaged carrier-frequency is used to compensate the current time record for carrier-frequency offsets from the center frequency.

Considerations When Using Auto Carrier Frequency

The following situations can bias the phase ramp estimation:

• Low frequency modulation, such as a periodic signal with fewer than 10 cycles over the time record.

• Phase discontinuities present in digital communication formats.

• Transients, such as carrier turn-on in the middle of the time record.

In cases where biases are unavoidable, auto carrier frequency should be turned off. Where possible, lock the analyzer's external reference to a reference coherent with the carrier to eliminate frequency errors.

Remote Command

Restriction and Notes

Example

Key Path

Mode

Preset

State Saved

[:SENSe]:ADEMod:CARRier:FREQuency:AUTO OFF|ON|0|1

[:SENSe]:ADEMod:CARRier:FREQuency:AUTO?

Greyed out if Demod Type is AM

ADEM:CARR:FREQ:AUTO ON

ADEM:CARR:FREQ:AUTO?

Meas Setup, Demod Setup

VSA

OFF

Saved in instrument state.

When Auto Carrier Frequency is on (for Demod types FM or PM), the calculated average carrier

328

Analog Demod

Meas Setup

frequency may be queried for any trace showing a demod result by using the following SCPI command.

Remote Command

Remote Command Notes

Example

Key Path

Mode

:DISPlay:ADEMod:TRACe[1]|2|3|4:CARRier:FREQuency?

This returns the result of the carrier frequency calculation (if Demod Mode is

FM or PM and Auto Carrier Freq is on) for the addressed trace (which must be assigned a demod result). Returns NaN otherwise.

DISP:ADEM:TRAC:CARR:FREQ?

SCPI only

VSA

Auto Carrier Phase

This allows you to turn on or off automatic carrier phase offset estimation for PM demodulation. Even with Auto Carrier Freq turned on, the PM demodulation may have a fixed or slowly varying phase offset.

This function estimates the phase offset and takes it out. A phase offset does not affect FM results, because the offset differentiates to zero.

Remote Command

Restriction and Notes

Example

Key Path

Mode

Preset

State Saved

[:SENSe]:ADEMod:CARRier:PHASe:AUTO OFF|ON|0|1

[:SENSe]:ADEMod:CARRier:PHASe:AUTO?

This is robust enough that it sometimes works even if Auto Carrier Freq is turned off, but it is recommended that you turn on Auto Carrier Freq along with this. Greyed out if Demod Type is AM or FM

ADEM:CARR:PHAS:AUTO ON

ADEM:CARR:PHAS:AUTO?

Meas Setup, Demod Setup

VSA

OFF

Saved in instrument state.

Meas Preset

For more information, see the section under the Preset key in the System Functions section.

Key Path

Mode

Meas Setup

VSA

329

Analog Demod

Mode

Mode

See

“Mode” on page 213 in the section "Common Measurement Functions" for a description of this

function.

330

Analog Demod

Mode Setup

Mode Setup

See “Mode Setup” on page 229 in the section "Common Measurement Functions" for a description of

this function.

331

Analog Demod

Single

Single

See

“Single (Single Measurement/Sweep)” on page 263 in the section "Common Measurement

Functions" for a description of this function.

332

Analog Demod

Trace/Detector

Trace/Detector

This section details the trace results accessible via the data key and also via SCPI, many of which are available in other measurements as well.

See “Trace/Detector” on page 428 in the section "Common Measurement Functions" for a description of

this function.

Select Trace

See “Select Trace” on page 428 in the section "Common Measurement Functions 2" for a description of

this function.

Data

This softkey brings up a menu of trace data choices for the selected trace. For more details and SCPI, see

Trace/Detector, “Data” on page 430 .

The following trace data results are available under the Pre-Demod submenu. They are identical to the results available in the Vector Analysis measurement.

Trace data

Spectrum

Inst Spectrum

Main Time

Inst Main Time

Gate Time

Raw Main Time

SCPI string form

"Spectrum1"

"Inst Spec1"

"Main Time1"

"Inst Main Time1"

"Gate Time1"

"Raw Main Time1"

"PSD1"

Description

Averaged result of successive Inst Spectrum results (If RMS or Max averaging is on). Otherwise, FFT of current windowed Main (or Gate)

Time

Instantaneous Spectrum is the FFT of the current windowed Main (or

Gate) Time. It is instantaneous in the sense that it isn't rms averaged, but it may included time-averaged data.

Same as Inst Main Time unless Time averaging is on, in which case it is the averaged result of successive Inst Main Time results.

If Time Gating is on, Main Time is the source to which the Gate is applied.

Instantaneous Main Time is the current corrected, resampled, time record.

Gate Time replaces Main Time as input to results if gating is on.

Time record as it comes from the hardware, before software resampling or corrections

Power spectrum divided by ResBW Power Spectral

Density (PSD)

CCDF "CCDF1"

CDF "CDF1"

Complementary Cumulative Distribution Function of all time date since last measurement restart

Cumulative Distribution Function of all time date since last measurement restart

333

Analog Demod

Trace/Detector

Trace data

PDF

SCPI string form

"PDF1"

Description

Probability Distribution Function of all time date since last measurement restart

Autocorrelation function of the current Main (or Gate) Time result Auto Correlation "Auto Correl1"

These Demodulation results are available under the Demod submenu. The name of the demod result and the trace title annotation are prefixed with the name of the demodulation type. For example, if you select

AM demodulation, then under Trace/Detector, Demod you will be able to select "AM Spectrum" and the trace will be labeled "Ch1 AM Spectrum".

Trace data

Spectrum

Inst Spectrum

Main Time

Inst Main Time

Gate Time

SCPI string form Description

"AnDemod Spectrum1" Same as calculation for input Spectrum, but using demod time data

"AnDemod Inst Spec1" Same as calculation for input Inst Spectrum, but using demod time data

"AnDemod Main

Time1"

"AnDemod Inst Main

Time1"

Main demodulated time record. Averaging and gating applied in the same manner as input Main Time.

Demodulated Inst Main Time

"AnDemod Gate

Time1"

"AnDemod PSD1"

Gated version of Demod Main Time, used as input to all other demod results if gating is on.

Power spectrum of demodulated signal divided by ResBW Power Spectral

Density (PSD)

CCDF

CDF

PDF

Auto Correlation

"AnDemod CCDF1"

"AnDemod CDF1"

"AnDemod PDF1"

"AnDemod Auto

Correl1"

Complementary Cumulative Distribution Function of all demodulated time date since last measurement restart

Cumulative Distribution Function of all demodulated time date since last measurement restart

Probability Distribution Function of all demodulated time date since last measurement restart

Autocorrelation function of the current demodulated Main (or Gate)

Time result

These data are also available (as they are in all measurements).

OBW Summary

Trace 1

ACP Summary

Trace 1

"OBW Summary Trc1" Table of Occupied Bandwidth results if OccBW is enabled on Trace 1 and Trace 1 has Spectrum or PSD data. Similar summaries are available for all traces.

"ACP Summary Trc1" Table of Adjacent Channel Power results if ACP is enabled on Trace

1 and Trace 1 has Spectrum or PSD data. Similar summaries are available for all traces.

334

Analog Demod

Trace/Detector

No Data "No Data" An empty trace

Format

This key provides a menu of Trace Formats from which to choose for the selected trace.

See Trace/Detector,

“Format” on page 443

in the section "Common Measurement Functions 2" for more information.

Digital Demod Trace Setup

This key provides a menu that allows you to set Digital Demod parameters for the selected trace.

See Trace/Detector, “Digital Demod Trace Setup” on page 444 in the section "Common Measurement

Functions 2" for more information.

Copy to Data Register

This key provides a menu of Data Registers numbered from 1 to 6 into which to copy the selected trace.

See Trace/Detector, “Copy to Data Register” on page 450

in the section "Common Measurement

Functions 2" for more information.

Phase/Delay Properties

This key provides a menu that allows you to set Phase and Delay properties for the selected trace.

See Trace/Detector, “Phase/Delay Properties” on page 451 in the section "Common Measurement

Functions 2" for more information.

ACP Setup

This key provides a menu that allows you to set ACP setup parameters for the selected trace.

See Trace/Detector, “ACP Setup” on page 453

in the section "Common Measurement Functions 2" for more information.

OBW Setup

This key provides a menu that allows you to set OBW setup parameters for the selected trace.

See Trace/Detector, “OBW Setup (Occupied Bandwidth)” on page 463

in the section "Common

Measurement Functions 2" for more information.

Trace Indicator Info

Pressing this key invokes the display of the currently selected trace indicator in the MSG bar at the bottom of the screen.

See Trace/Detector,

“Trace Indicator Info” on page 469

in the section "Common Measurement Functions

2" for more information.

335

Analog Demod

Trigger

Trigger

Triggering is used to determine when a measurement should start taking data. There are several available trigger sources. For each trigger source, there are associated setup parameters. Typically, a trigger event is generated when a signal (or a characteristic of the signal) crosses a defined trigger level (or threshold) on a rising or falling slope. The measurement begins at a specified time delay from the trigger point. The delay may be negative, allowing pretrigger data to be taken. Each trigger source has associated its own trigger level, slope, and delay settings.

See

“Trigger” on page 470

in the section "Common Measurement Functions 2" for a description of this function.

336

Analog Demod

View/Display (View Presets)

View/Display (View Presets)

The View/Display key displays a menu that enables you to select display parameters for the current measurement.

View Presets affect the trace layout, trace data assignment, scaling and formatting but do not affect hardware measurement setup.

Remote Command

Example

Key Path

Mode

:DISPlay:ADEMod:VIEW:PRESet DSPectrum|STATistics

DISP:ADEM:VIEW:PRES DSP

View/Display

VSA

Display

See “Display” on page 349 in the section "Common Measurement Functions" for a description of this

function.

Layout

See “Layout” on page 479

in the section "Common Measurement Functions 2" for a description of this function.

Preset View: Demod Spectrum/Time

This preset shows a quad layout with the Demod Spectrum in trace 1, Demod Main Time in trace 2, the input Spectrum in trace 3, and the input Main Time in trace 4.

Key Path

Mode

View/Trace

VSA

Preset View: Statistics

This preset shows a stacked 2 layout with the CCDF of the input in trace 1 and input Main Time in trace

2.

Key Path

Mode

View/Trace

VSA

337

Analog Demod

View/Display (View Presets)

338

9

Digital Demod

The Digital Demod measurement is accessed from the Meas hardkey. The Digital Demod measurement builds upon basic Vector analysis by including flexible demodulation of a wide variety of standard and custom single-carrier modulation formats. You may simultaneously view pre-demod time and spectrum displays, demodulated signal, reconstructed reference signal, recovered symbols and various error traces and summaries.

Measurements are possible on continuous or pulsed (burst) carriers (such as TDMA). In addition, you can specify a sync pattern and an offset to look at selected segments of demodulated data.

The digital demodulator uses your signal to generate an ideal signal (called I/Q reference or FSK reference). You can compare the measured signal to the reference signal to quantify and locate errors in your signal.

Digital demodulation has built-in filters which may be applied to the measured signal as well as to the reference signal. This allows you maximum flexibility in comparing your signal to an ideal signal.

Additionally, this allows complete flexibility to probe any analog point in a communication system.

This topic contains the following sections:

“Remote Command Results for Digital Demod Measurement” on page 339

“Front Panel Results” on page 342

Remote Command Results for Digital Demod Measurement

The Digital Demod measurement is invoked remotely by the following:

:CONFigure:DDEMod

:CONFigure:DDEMod:NDEFault

:INITiate:DDEMod

All trace and tabular data results are available using CALCulate:DDEMod:DATA commands. These commands also enable you to get names and units of results.

Symbol/Error Table results may also be obtained using the FETCh or READ commands. The Sym/Err table must be assigned to a trace in order to obtain valid results.

:FETCh:DDEMod[n]?

:READ:DDEMod[n]?

NOTE

The MEASure:DDEMod? command is not supported by the Digital Demod measurement.

For more information and remote commands, see “Remote SCPI Commands and Data Queries” on page 482

in the section, “Common Measurement Functions 2”.

Also see Trace/Detector,

“Data” on page 430

in the section “Common Measurement Functions 2” for

339

Digital Demod more measurement SCPI commands.

340

Digital Demod

Condition N

All Mod

Formats

Results Returned

Not specifie d, or n=1

Error Summary Table

Returns 30 comma-separated scalar results, corresponding to the items in the table portion of the Syms/Err trace. Note some values are not available (n/a) for some formats.NaN

(9.91 E 37) is returned for results that are not available.

1. EVM rms (% rms) (n/a for FSK)

2. EVM peak (% pk) (n/a for FSK)

3. symbol position of EVM peak (n/a for FSK)

4. offset EVM rms (% rms) (OQPSK only, n/a otherwise)

5. offset EVM peak (% pk) (OQPSK only, n/a otherwise)

6. symbol position of Offset EVM peak (OQPSK only, n/a otherwise)

7. FSK err rms (% rms) (FSK only, n/a otherwise)

8. FSK err peak (% pk) (FSK only, n/a otherwise)

9. symbol position of FSK err peak (FSK only, n/a otherwise)

10. magnitude error rms (% rms).

11. magnitude error peak (% pk)

12. symbol position of magnitude error peak

13. phase error rms (deg) (n/a for FSK)

14. phase error peak (deg pk) (n/a for FSK)

15. symbol position of phase error peak (n/a for FSK)

16. frequency error (Hz) (n/a for FSK)

17. carrier offset (Hz) (FSK only, n/a otherwise)

18. SNR(MER) (dB) (QPSK, QAM, APSK and VSB only, n/a otherwise)

19. FSK deviation (Hz) (FSK only, n/a otherwise)

20. Pilot Level (dB) (8 VSB only, n/a otherwise)

21. time offset (s) (triggered APSK only, n/a otherwise)

22. IQ offset (dB) (n/a for FSK, VSB)

23. amplitude droop (dB/sym) (n/a for QPSK, OQPSK, MSK Type 1, QAM, APSK, VSB and FSK)

24. rho (QPSK and OQPSK only, n/a otherwise)

25. quadrature error (deg) (n/a for BPSK, VSB and FSK)

26. gain imbalance (dB) (n/a for BPSK, VSB and FSK)

27. R2/R1 ratio (dimensionless) (APSK only, n/a otherwise)

28. R3/R1 ratio (dimensionless) (APSK 32 only, n/a otherwise)

29. peak EVM mean (%) (EDGE only, n/a otherwise)

30. 95% EVM (%) (EDGE only, n/a otherwise)

341

Digital Demod

Also available are the standard ACP and OBW tables.

See

“Trace/Detector” on page 383

.

Front Panel Results

Digital Demod results may be displayed in any trace, and the traces viewed in a variety of layouts that show 1, 2, 3, or 4 traces at a time. Each trace may be scaled as desired regardless of measurement settings, or auto-scaled to reflect measurement settings. Data may be formatted in a variety of ways. (For example, you may view the log magnitude of complex data, the real or imaginary part, etc.) You may use

View Presets to view frequently used results, or to provide a familiar starting point from which you may customize your own view.

Key Path

Mode

Front Panel

VSA

342

Digital Demod

AMPTD Y Scale

AMPTD Y Scale

See “AMPTD Y Scale (Amplitude)” on page 361 in the section "Common Measurement Functions 2"

for for a description of this function.

343

Digital Demod

Auto Couple

Auto Couple

The Auto Couple key forces all Auto/Man functions into Auto. These include the following functions:

• Frequency Step

• X Scale

• Y Axis Unit Preference

• Frequency Points

• Playback Start/Stop

NOTE

This function does not invoke the Auto tune function, nor does it cause any Y autoscaling or Input Auto ranging.

For more information, see

“Auto Couple” on page 33

in the section "Common Measurement Functions".

Key Path

Mode

Front Panel

VSA

344

Digital Demod

BW

BW

See “BW (Bandwidth)” on page 368 in the section "Common Measurement Functions 2" for a

description of this function.

345

Digital Demod

Cont (Continuous)

Cont (Continuous)

See

“Cont (Continuous Measurement/Sweep)” on page 35

in the section "Common Measurement

Functions 1" for a description of this function.

FFT Window

See “FFT Window” on page 371 in the section "Common Measurement Functions 2" for a description of

this function.

346

Digital Demod

FREQ Channel

FREQ Channel

See “FREQ Channel” on page 371

in the section "Common Measurement Functions 2" for a description of this function.

347

Digital Demod

Input/Output

Input/Output

See

“Input/Output” on page 43

in the section "Common Measurement Functions " for a description of this function.

348

Digital Demod

Marker

Marker

See “Marker” on page 374

in the section "Common Measurement Functions 2" for a description of this function.

349

Digital Demod

Marker Function

Marker Function

See

“Marker Function” on page 392

in the section "Common Measurement Functions 2" for a description of this function.

350

Digital Demod

Marker To

Marker To

See “Marker -> (Marker To)” on page 390 in the section "Common Measurement Functions 1" for a

description of this function.

351

Digital Demod

Meas (Measure)

Meas (Measure)

See

“Meas” on page 155 in the section "Common Measurement Functions " for a description of this

function.

352

Digital Demod

Meas Setup

Meas Setup

The common Meas Setup functions are documented in

“Meas Setup” on page 400

in the section

"Common Measurement Functions 2".

Avg Number

See “Avg Number” on page 400 in the section "Common Measurement Functions 2" for a description of

this function.

Average Mode

See “Average Mode” on page 401 in the section "Common Measurement Functions 2" for a description

of this function.

Average Setup

See “Average Setup” on page 402 in the section "Common Measurement Functions 2" for a description

of this function.

Demod Setup

This key displays a menu that allows you to adjust digital demodulation parameters.

Key Path

Mode

Meas Setup

VSA

Modulation Format

This allows you to select the digital communication format that is used by the demodulator. The selection includes:

• QAM formats: QAM 16, QAM 32, QAM 64, QAM 128, QAM 256, QAM 512, and QAM 1024

• PSK formats: BPSK, QPSK, Offset QPSK, /4 DQPSK, DQPSK, 8PSK, /8 D8PSK, and D8PSK

• MSK type 1 and type 2

• FSK formats: FSK 2, FSK 4, FSK 8, and FSK 16

• DVB QAM formats: DVB QAM 16, DVB QAM 32, DVB QAM 64, DVB QAM 128, and DVB

QAM 256

• VSB formats: VSB 8 and VSB 16

• APSK (amplitude/phase shift keying) formats: APSK 16, APSK 16 w/DVB, APSK 32, APSK 32 w/DVB

• EDGE

353

Digital Demod

Meas Setup

Custom State Definitions

The states that correspond to each constellation position for each format are defined in the following file:

D:\User_My_Documents\Instrument\My Documents\VXA\Data\Constellation\Current.csd

If you want to define custom states, edit this file using Notepad. The file contains instructions and examples to guide you. After saving your changes, you must exit and restart the XSA program for the changes to take effect.

To return the original state definitions, simply delete this file and restart XSA. A new Current.csd file will be created with the original state definitions restored.

Remote Command

Example

Key Path

Mode

Preset

State Saved

Range

[:SENSe]:DDEMod:MODulation

BPSK|QPSK|OQPSK|PI4DQPSK|DQPSK|PSK8|PI8DPSK8|DPSK8|QAM1

6|QAM32|QAM64|QAM128|QAM256|QAM512|QAM1024|FSK2|FSK4|FS

K8|FSK16|MSK1|MSK2|EDGE|APSK16|APSK32|DVBAPSK16|DVBAPSK

32|DVBQAM16|DVBQAM32|DVBQAM64|DVBQAM128|DVBQAM256|VSB8|

VSB16

[:SENSe]:DDEMod:MODulation?

DDEM:MOD QPSK

DDEM:MOD?

Meas Setup, Demod Setup

VSA

QPSK

Saved in instrument state.

BPSK | QPSK | Offset QPSK | /4DQPSK | DQPSK | 8PSK | /8 DPSK8 |

D8PSK | QAM 16 | QAM 32 | QAM 64 | QAM 128 | QAM 256 | QAM 512 |

QAM 1024 | FSK 2 | FSK 4 | FSK 8 | FSK 16 | MSK Type 1 | MSK Type 2 |

EDGE | APSK 16 | APSK 32 | APSK 16 w/DVB | APSK 32 w/DVB | DVB

QAM 16 | DVB QAM 32 | DVB QAM 64 | DVB QAM 128 | DVB QAM 256

| VSB 8 | VSB 16

Symbol Rate

The Symbol Rate key allows you to set the symbol rate (symbols per second) for the analyzer's digital demodulator. Set this parameter to match the symbol rate of your system.

In digital modulation, the symbol rate determines the rate (frequency) at which symbols occur. A symbol may consist of one or more bits as determined by the modulation format. For example, in a BPSK system, each symbol represents 1 bit; in a QPSK system, each symbol represents 2 bits.

Symbols are valid only at the timing instant when the receiver interprets the signal. This timing instant is called the detection-decision point.

The analyzer's demodulator uses the symbol rate determine the frequency of your detection-decision points. It is important that you set the symbol rate to match exactly the symbol rate of your system, because the symbol clock frequency is not estimated.

Note that the more complex your modulation format, the more critical it is that the symbol rate be exact.

354

Digital Demod

Meas Setup

Specifying an incorrect symbol rate introduces errors into the demodulation process.

The analyzer can accurately measure symbol rates that are less than the maximum span of the analyzer.

There may be instances where you want to set the symbol rate beyond what can be measured. The analyzer lets you do this, but accuracy is not specified

If you enter a symbol rate that is slightly different than the symbol rate of your signal, the EVM (error vector magnitude) is typically small at the center of the result length and increases linearly towards the ends of the result length.

The symbol rate determines the maximum frequency span (information bandwidth) that you can measure. For QAM and PSK signals, the symbol rate also determines the minimum frequency span that meets published specifications.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:DDEMod:SRATe <frequency>

[:SENSe]:DDEMod:SRATe?

Span is forced to be < 15.625 * (Symbol rate)

DDEM:SRAT 1 MHZ

DDEM:SRAT?

Meas Setup, Demod Setup

VSA

3840000 HZ

Saved in instrument state.

1

9.9E+37

Meas Interval

This key sets the number of symbols that the demodulation will analyze. This and the symbol rate set the overall time record length (in seconds) that is used by the demodulator. It also sets indirectly sets the resolution bandwidth for the various spectrum results. (The ResBW cannot be set independently.)

The resolution bandwidth and Time length are related by the following equation:

Res BW = ENBW / T where:

ENBW is the normalized effective noise bandwidth of the Window (see the FFT Window topic for more details).

For the pre-demod Spectrum result, T = 1.2*(Meas Interval)/Symbol Rate.

For the all other Spectrum results, T = (Meas Interval)/Symbol Rate.

The resolution bandwidth is annotated below any spectrum trace. To programmatically query the resolution bandwidth, use the following (with the trace number for <n>)

355

Digital Demod

Meas Setup

CALC:DDEM:DATA<n>:HEAD? "ResBW"

Remote Command

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:DDEMod:SWEep:POINts <integer>

[:SENSe]:DDEMod:SWEep:POINts?

DDEM:SWE:POIN 137

DDEM:SWE:POIN?

Meas Setup, Demod Setup

VSA

200

Saved in instrument state.

10

4096

Points / Symbol

Allows you to set how many points are displayed per symbol in time displays of demodulated data. The allowed values are 1, 2, 4, 5, 10 and 20.

Points/Symbol do not apply to W-CDMA(3GPP) or cdma2000 demodulation.

Minimum (except OQPSK):

Minimum (OQPSK):

Maximum:

EDGE default (see below):

1 point per symbol

2 points per symbol

20 points per symbol

1 point per symbol

For example, if the value of Points/Symbol is 1, each display point corresponds to a symbol. If the value is 5, the 5th display point corresponds to a symbolin this case, an IQ diagram would show 4 display points between each symbol.

Saving Points/Symbol: When you save a digitally demodulated trace, the Points/Symbol value used to create the trace is saved with the trace. When you recall the trace, the analyzer displays the trace with the

Points/Symbol value used to create the tracethe analyzer DOES NOT use the current value of

Points/Symbol.

MSK Demodulation: For the MSK demodulation format, changing Points/Symbol affects the error data displayed in the Alphabetical Listing of Error Summary Data. This occurs because the analyzer uses all points to compute Error data results for MSK, whereas the analyzer uses only the points that occur at the symbol times to compute error data results for other demodulation formats.

OQPSK Demodulation: For OQPSK, an even number of Points/Symbol are required due to the offset between I and Q. If you specify an odd value for Points/Symbol, the analyzer chooses the next, lower, even value.

EDGE demodulation: For `EDGE` demodulation format and the Points/Symbol is set to 1 (default), the

IQ Meas Time, IQ Magnitude Error, IQ Phase Error and Error Summary Table trace data results are the

356

Digital Demod

Meas Setup

ISI (inter-symbol interference) compensated values. That is, when the points/symbol is set to 1 (default), the analyzer removes the effects of ISI (inter-symbol interference), which provides a "clean" IQ Meas

Time constellation diagram. For points/symbol greater than one, the trace data results are not compensated for the effects of ISI. For values greater than 1 point/symbol, the symbols in EDGE constellation diagrams may appear randomly placed due to the effects of ISI.

Couple to Gain Imb./Quad Skew: This measurement parameter selection controls the number of points per symbol used to calculate the IQ Gain Imbalance and Quadrature Skew symbol error data results. For further information, see Couple to Gain Imb./Quad Skew, IQ Gain Imbalance, and Quadrature Skew.

Remote Command

Example

Key Path

Mode

Notes

Preset

State Saved

Min

Max

[:SENSe]:DDEMod:PPSYmbol <integer>

[:SENSe]:DDEMod:PPSYmbol?

DDEM:PPSY 2

DDEM:PPSY?

Meas Setup, Demod Setup

VSA

The only supported values are 1, 2, 4, 5, 10 and 20. Numeric entries are rounded to the nearest valid value.

5

Saved in instrument state.

1

20

Gain Imb/Quad Skew Coupling

Controls what measurement data is included in the Quadrature Skew Error and IQ Gain Imbalance error data calculations.

• Off: Calculations use one Point per Symbol.

• On: Calculations use the value shown in the Points per Symbol parameter box.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

:CALCulate:DDEMod:PPSYmbol:COUPle OFF|ON|0|1

:CALCulate:DDEMod:PPSYmbol:COUPle?

No

CALC:DDEM:PPSY:COUP OFF

CALC:DDEM:PPSY:COUP?

Meas Setup, Demod Setup

VSA

OFF

Saved in instrument state.

357

Digital Demod

Meas Setup

Meas Filter

Allows you to select the Meas Filter used by the receiver.

Data filtering is used in digital demodulation to limit bandwidth and reduce intersymbol interference.

This analyzer includes several commonly used filter types and has the ability to apply user-defined filters. All filters are computed to 20 symbols in length. If the filter alpha is < 0.2, the analyzer uses 40 symbols to compute filter length (for most formats).

The shape and width of a filter is defined by the alpha (for cosine filters) or the BT (for Gaussian filters).

The alpha or BT indicates the filter roll-off (or excess bandwidth) of the selected filter which occurs due to the practical inability of filter technology to build a perfectly square (brick-wall) filter which would have an alpha of 0 (no excess bandwidth). For example, a typical filter with an alpha of 0.3 has a bandwidth 30% greater than the theoretical minimum.

The analyzer's digital demodulator produces two signals: a measured and a reference signal. These signals are called I/Q Measured and I/Q Reference or, for FSK measurements, FSK Measured and FSK

Reference.

You can select different filters for the measured and reference signals, as shown in the following, generic block diagram.

Note that for FSK signals, filtering is baseband and occurs after the FM demodulator.

The measured signal is the signal that results after demodulating your waveform. The reference signal is the signal that would result after demodulating your signal if your signal were ideal (contained no errors).

Notice that there are separate filters for the measured and reference signals. You MUST select the correct filter for both signals.

Filtering for various communication systems may occur either at the transmitter or the receiver; or the filtering may be distributed between the transmitter and the receiver. This is an important concept which affects your filter selection for the measured and reference signals. The analyzer's measured filter represents filtering in the system's receiver while the reference filter represents filtering in the entire system. Both filters share the same alpha/BT. The following table shows some examples of filter selection:

If the transmitter filter is:

root raised cosine

Raised cosine

Gaussian any type

The measure filter should be:

root raised cosine none none user defined

The reference filter should be:

raised cosine raised cosine

Gaussian any type

You can modify the shape and width of Gaussian and Nyquist (cosine) filters. The shape and width is defined by the alpha (for cosine filters) or the BT (for Gaussian filters). All filters are computed to 20 symbols in length. If the filter alpha is < 0.2, the analyzer uses 40 symbols to compute filter length (for

358

Digital Demod

Meas Setup

most formats).

Remote Command

Example

Key Path

Mode

Notes

Preset

State Saved

Range

Readback Text

[:SENSe]:DDEMod:FILTer:MEASurement

NONE|RRCosine|GAUSsian|EDGE|IS95EQ|RECTangle|LPASs|USER

[:SENSe]:DDEMod:FILTer:MEASurement?

DDEM:FILT:MEAS RRC

DDEM:FILT:MEAS?

Meas Setup, Demod Setup

VSA

USER defined requires you to specify a register (see Meas User Defined).

RRC

Saved in instrument state.

No Filter | Root Raised Cosine | Gaussian | EDGE | CDMA (IS–95A Base EQ)

| Rectangular | Low Pass | User Defined

No Filter | RRC | Gaussian | EDGE | CDMA (EQ) | Rectangular | Low Pass |

User Defined

Meas User Defined

Pressing this key sets the Meas Filter to User Defined. Pressing it a second time allows you to select a data register which contains data that defines the filter.

The analyzer lets you use a filter of your design for the I/Q measured filter or the I/Q reference filter.

Typical applications for user-defined filters include:

• Custom filters other than those provided. In this case both measured and reference filters are user defined.

• Modified filters that are based on the raised cosine filter but add channel equalization. In this case, the measure filter is a user-defined filter and the reference filter is standard.

To define a filter, create a trace file containing the time-domain impulse response of the filter. (The easiest way to start is to export a time domain trace into a file in text or csv format and modify it. See

Utility (System Functions), Save, Export Trace Data .) Then recall the trace file into the desired register.

(See Utility (System Functions), Recall, Import Trace Data.) The data must satisfy these criteria:

• Only be real data, not complex

• Only be time-domain data. If the data domain type cannot be determined from the trace file

(unknown domain), the analyzer will default to time-domain data. If any other domain is used, for example the frequency-domain, the analyzer will reject the user-defined trace file and use a root raised cosine filter as the Measurement filter and a raised cosine filter as the Reference filter.

• Contain > 0 and <= 20 symbols (401 samples) in overall length. However, if the filter alpha is < 0.2, the trace length must be <= 40 symbols (801 samples) in overall length.

• Contain 20 samples-per-symbol. For example, 401 samples = 20 symbols at 20 points-per-symbol.

359

Digital Demod

Meas Setup

• Use an odd number of points so that the center of the impulse is positioned on a symbol, this would be the 201st point in a 401 point trace. The middle sample is assumed to be at t = 0.

If the trace file does not satisfy all of previously mentioned criterion, the analyzer rejects the user-defined trace file and defaults to using the root raised cosine filter as the Measurement filter and the raised cosine filter as the Reference filter.

Accuracy of user-defined filters is undefined. In addition, the value of Alpha/BT has no effect on user-defined filters.

Remote Command

Coupling

Example

Key Path

Mode

Notes

Preset

State Saved

Range

[:SENSe]:DDEMod:FILTer:MEASurement:REGister

D1|D2|D3|D4|D5|D6

[:SENSe]:DDEMod:FILTer:MEASurement:REGister?

Filter Alpha influences length constraint (see above) but has no other effect.

DDEM:FILT:MEAS:REG D1

DDEM:FILT:MEAS:REG?

Meas Setup, Demod Setup, Meas Filter

VSA

Individual Data register selections are greyed out if they do not contain appropriate data for use as filter coefficients.

D1

Saved in instrument state.

Data 1 | Data 2 | Data 3 | Data 4 | Data 5 | Data 6

Ref Filter

Pressing this key selects the Ref Filter that represents the cascaded transmit and receive filter.

This analyzer includes several commonly used filter types and has the ability to apply user-defined filters. All filters are computed to 20 symbols in length. If the filter alpha is < 0.2, the analyzer uses 40 symbols to compute filter length (for most formats).

The shape and width of a filter is defined by the alpha (for cosine filters) or the BT (for Gaussian filters).

The alpha or BT indicates the filter roll-off (or excess bandwidth) of the selected filter which occurs due to the practical inability of filter technology to build a perfectly square (brick-wall) filter which would have an alpha of 0 (no excess bandwidth). For example, a typical filter with an alpha of 0.3 has a bandwidth 30% greater than the theoretical minimum.

The analyzer's digital demodulator produces two signals: a measured and a reference signal. These signals are called I/Q Measured and I/Q Reference or, for FSK measurements, FSK Measured and FSK

Reference. The following table shows some examples of filter selection:

If the transmitter filter is:

root raised cosine

The measure filter should be:

root raised cosine

The reference filter should be:

raised cosine

360

Digital Demod

Meas Setup

If the transmitter filter is:

raised cosine

Gaussian any type

The measure filter should be:

none none user defined

The reference filter should be:

raised cosine

Gaussian any type

You can modify the shape and width of Gaussian and Nyquist (cosine) filters. The shape and width is defined by the alpha (for cosine filters) or the BT (for Gaussian filters). All filters are computed to 20 symbols in length. If the filter alpha is < 0.2, the analyzer uses 40 symbols to compute filter length (for most formats).

Matched filtering of a demodulated signal is not available with the MSK demodulators. The measured filter is normally off. A user defined filter can be selected. Its primary use is for additional band-limiting and channel equalization.

Remote Command

Example

Key Path

Mode

Notes

Preset

State Saved

Range

Readback Text

[:SENSe]:DDEMod:FILTer:REFerence

RCOSine|RRCosine|GAUSsian|EDGE|IS95BB|RECTangle|HSINe|U

SER

[:SENSe]:DDEMod:FILTer:REFerence?

DDEM:FILT:REF RCOS

DDEM:FILT:REF?

Meas Setup, Demod Setup

VSA

Individual Data register selections are greyed out if they do not contain appropriate data for use as filter coefficients. USER defined requires you to specify a register (see Ref User Defined).

RCOS

Saved in instrument state.

Raised Cosine | Root Raised Cosine | Gaussian | EDGE | CDMA (IS–95 Base)

| Rectangular | Half Sine | User Defined

Raised Cosine | RRC | Gaussian | EDGE | CDMA (Base) | Rectangular | Half

Sine | User Defined

Ref User Defined

Pressing this key causes the Ref Filter to become User Defined. Pressing it a second time displays a menu that allows you to select the data register which containing the data that defines the filter.

To define a filter, create a trace file containing the impulse response of the filter. (The easiest way to start is to export a time domain trace into a file in text or CSV format and modify it Then recall the file into the desired register. The data must satisfy these criteria:

• Only be real data, not complex

361

Digital Demod

Meas Setup

• Only be time-domain data. If the data domain type cannot be determined from the trace file

(unknown domain), the analyzer will default to time-domain data. If any other domain is used, for example the frequency-domain, the analyzer will reject the user-defined trace file and use a root raised cosine filter as the Measurement filter and a raised cosine filter as the Reference filter.

• Contain > 0 and <= 20 symbols (401 samples) in overall length. However, if the filter alpha is < 0.2, the trace length must be <= 40 symbols (801 samples) in overall length.

• Contain 20 samples-per-symbol. For example, 401 samples = 20 symbols at 20 points-per-symbol.

• Use an odd number of points so that the center of the impulse is positioned on a symbol, this would be the 201st point in a 401 point trace. The middle sample is assumed to be at t = 0.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Range

[:SENSe]:DDEMod:FILTer:REFerence:REGister

D1|D2|D3|D4|D5|D6

[:SENSe]:DDEMod:FILTer:REFerence:REGister?

Filter Alpha influences length constraint (see above) but has no other effect.

DDEM:FILT:REF:REG D1

DDEM:FILT:REF:REG?

Meas Setup, Demod Setup, Ref Filter

VSA

D1

Saved in instrument state.

Data 1 | Data 2 | Data 3 | Data 4 | Data 5 | Data 6

Alpha / BT

Determines the filter characteristics of the Raised cosine, Root-raised cosine and Gaussian filters used by the analyzer's digital demodulator. These characteristics apply to both the Meas and Ref filters.

Allowable values,

Raised cosine, root-raised cosine filters: .05 to 1

Gaussian filters: .05 to 100

Remote Command

Coupling

Example

Key Path

Mode

[:SENSe]:DDEMod:ALPHa <real>

[:SENSe]:DDEMod:ALPHa?

Alpha also determines length criteria for user defined meas and ref filters

DDEM:ALPH 0.22

DDEM:ALPH?

Meas Setup, Demod Setup

VSA

362

Digital Demod

Meas Setup

Preset

State Saved

Min

Max

0.22

Saved in instrument state.

0.05

100

Burst/Sync Search

This key displays a menu that allows you to select Burst and Sync searches, and to select search parameters for the current measurement..

Key Path

Mode

Meas Setup, Demod Setup

VSA

Search Length

This defines the time length over which the analyzer will search for a burst and/or sync word. . You can specify search length in number of symbols or units of time.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:DDEMod:SYNC:SLENgth <time>

[:SENSe]:DDEMod:SYNC:SLENgth?

Minimum: Meas Interval / Symbol Rate

Maximum: Depends on span

DDEM:SYNC:SLEN 200 US

DDEM:SYNC:SLEN?

Meas Setup, Demod Setup, Burst/Sync Search

VSA

666.666667E–6

Saved in instrument state.

0

9.9E+37

Burst Search

This key turns on or off burst search

Remote Command

Example

[:SENSe]:DDEMod:SYNC:BURSt:STATe OFF|ON|0|1

[:SENSe]:DDEMod:SYNC:BURSt:STATe?

DDEM:SYNC:BURS:STAT OFF

DDEM:SYNC:BURS:STAT?

363

Digital Demod

Meas Setup

Key Path

Mode

Preset

State Saved

Meas Setup, Demod Setup, Burst/Sync Search

VSA

OFF

Saved in instrument state.

Sync Search

This key turns on or off Sync Word search. Sync search lets you use a synchronization pattern to isolate a segment of your signal for display and analysis. The analyzer searches through demodulated data to find your sync pattern, and then uses the Result Length to determine how much data to display and the

Search Offset to display data relative to the sync pattern.

NOTE

The sync pattern must be a multiple of the number of bits-per-symbol. For example, if the number of bits-per-symbol is 4 (as with 16 QAM), then the number of bits in the sync pattern must be a multiple of four. Sync search lets you specify any number of bits for the sync pattern; however, bits that are not a multiple of the bits-per-symbol are truncated.

Remote Command

Example

Key Path

Mode

Preset

State Saved

[:SENSe]:DDEMod:SYNC:SWORd:STATe OFF|ON|0|1

[:SENSe]:DDEMod:SYNC:SWORd:STATe?

DDEM:SYNC:SWOR:STAT OFF

DDEM:SYNC:SWOR:STAT?

Meas Setup, Demod Setup, Burst/Sync Search

VSA

OFF

Saved in instrument state.

Sync Pattern

This allows you to define a bit pattern for the sync word search. The maximum length of the pattern is 32 symbols. When you press this key, an editor dialogappears that allows you to define the pattern in binary or hexadecimal.

The analyzer can only search for search patterns that are a multiple of the number of bits-per-symbol.

The analyzer assembles the bits in the search pattern into an integer number of symbols, ignoring any trailing bits that cannot complete a full symbol.

For example, if the number of bits-per-symbol is 4 (as with 16 QAM), and your search pattern contains

18 bits, the analyzer only uses the first 16 bits during sync search and ignores the last two bits.

Although the underlying sync pattern is binary, the editor allows you to enter bit patterns in either binary or hexadecimal.

364

Digital Demod

Meas Setup

Hexadecimal entries are in symbol table hex format. In this format, each symbol is represented by 1 to 3 hex digits. The least significant bits are used make up the symbol and any extra bits are ignored. Thus, for 6 bits per symbol, the binary value of 101101 is encoded in hex as 2D. If you entered AD as the symbol value, it would be converted to the same binary value as 2D would, because the upper 2 bits would be ignored. 3-bit-wide symbols are represented by 1 hex digit with the most significant bit ignored.

Predefined sync patterns are available for GSM/EDGE and APSK. If you choose one of these formats from the list, it populates the sync pattern hex and binary text boxes with their predefined value.

When Separate Symbols is turned on , spaces appear between symbols in the binary representation.

You can navigate around the dialog without needing a mouse. Pressing tab right and tab left move you from one control to the next. Arrow keys and the knob can be used to navigate within a list or editing box. The space key will toggle the Separate Symbols check box on and off. Softkeys and the numeric entry keys are used to enter patterns. Press Select or Done to complete the entry. Cancel or Return returns exits the dialog without changing the pattern.

Remote Command

[:SENSe]:DDEMod:SYNC:SWORd:PATTern <string>

[:SENSe]:DDEMod:SYNC:SWORd:PATTern?

365

Digital Demod

Meas Setup

Example

Key Path

Mode

Notes

DDEM:SYNC:SWOR:PATT '1011010'

DDEM:SYNC:SWOR:PATT?

Meas Setup, Demod Setup, Burst/Sync Search

VSA

<string> must be a string of 1s and 0s only. The maximum string length is 320 bits.

=""

Saved in instrument state.

Preset

State Saved

Sync Offset

This key specifies the time (in symbols) between the start of the measurement data and the start of the sync word. If positive, the sync word starts after the start of the measurement data. If negative, the sync word starts before the start of the measurement data.

The minimum and maximum offsets you can enter depend on these parameters:

Search Length

Result Length

Sync Pattern

Basically, you can enter any offset such that the result length falls within the search length. Increasing any of these parameters affects the maximum positive or negative offset that you can enter as follows:

Increasing search length increases the maximum positive or negative offset that you can enter.

Increasing result length decreases the maximum negative offset that you can enter but has no effect on the maximum positive offset.

Increasing the length of the sync pattern decreases the maximum positive offset that you can enter but has no effect on the maximum negative offset.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

[:SENSe]:DDEMod:SYNC:SWORd:OFFSet <integer>

[:SENSe]:DDEMod:SYNC:SWORd:OFFSet?

Max and min constrained by Search Length, Sync Pattern length

DDEM:SYNC:SWOR:OFFS –3

DDEM:SYNC:SWOR:OFFS?

Meas Setup, Demod Setup, Burst/Sync Search

VSA

0

Saved in instrument state.

see coupling

366

Digital Demod

Meas Setup

Max see coupling

Advanced Dig Demod

This key displays a menu that allows you to select advanced demodulation parameters for the current measurement. These settings are for advanced users and do not normally require adjustment for most common measurements.

Key Path

Mode

Meas Setup, Demod Setup

VSA

Clock Adjust

This key allows you to adjust symbol clock timing in fractions of a symbol. The adjustment is relative to the symbol clock time that is computed by the demodulation algorithm. Some digital communications systems contain nonlinearities that can bias the digital demodulator's estimation of the symbol clock position. You can use clock adjust to compensate for this offset and obtain a lower EVM (Error Vector

Magnitude).

Specifying a clock adjust only affects the I/Q measured trace. It does not affect the I/Q reference trace.

Use the eye diagram with an eye length of one (1) to observe the accuracy of the symbol clock timing.

You may also want to monitor the EVM (Error Vector Magnitude) in the symbol table summary while adjusting clock adjust to obtain the optimum symbol timing.

Clock adjust is reset to 0.0 on power-up or when you select Preset.

Remote Command

Example

Key Path

Mode

Notes

Preset

State Saved

Min

Max

[:SENSe]:DDEMod:CADJust <real>

[:SENSe]:DDEMod:CADJust?

DDEM:CADJ 0.14

DDEM:CADJ?

Meas Setup, Advanced Dig Demod

VSA

The parameter is interpreted as a fraction of a sample

0

Saved in instrument state.

–0.5

0.5

IQ Rotation

This rotates the Meas/Ref Time data and corresponding ideal state positions by a user defined amount ranging from –360 degrees to 360 degrees. The Rotation parameter affects the IQ Gain Imbalance and

367

Digital Demod

Meas Setup

Quadrature Skew error data results.

Remote Command

Example

Key Path

Mode

Notes

Preset

State Saved

Min

Max

:CALCulate:DDEMod:IQRotation <real>

:CALCulate:DDEMod:IQRotation?

CALC:DDEM:IQR 45

CALC:DDEM:IQR?

Meas Setup, Advanced Dig Demod

VSA

The numeric parameter is interpreted as degrees.

0

Saved in instrument state.

–360

360

IQ Normalize

Turns IQ Normalize on and off. When IQ Normalize is on, the Meas Time and Ref Time data is normalized so the extreme points have a value of 1. For quadrature modulation types, the outermost points of the constellation are normalized to 1. (Note, for non-square QAM constellation, the points that appear at the corners of the containing square are normalized to 1.) For FSK constellations, the deviation is normalized to 1.

When IQ Normalize is turned off, the actual data values based on the input signal level are plotted on the constellation.

When normalization is ON, the analyzer normalizes or scales the demodulated trace data results to a nominal value of 1. Normalization is performed on these traces:

IQ measured time for Digital, WLAN-OFDM, WLAN-DSSS/CCK/PBSS, WCDMA, cdma2000,

TD-SCDMA and 1xEV-DO demodulation.

IQ reference time for Digital, WLAN-OFDM, WLAN-DSSS/CCK/PBSS, WCDMA, cdma2000,

TD-SCDMA and 1xEV-DO demodulation.

Error vector time

FSK measured time (FSK measurements)

FSK measured reference (FSK measurements)

Carrier error magnitude (FSK measurements)

FSK error (FSK measurements)

CDP and CDE measurements for W-CDMA, cdma2000 1xEV-DO and TD-SCDMA demodulation)

Remote Command

:CALCulate:DDEMod:NORMalize OFF|ON|0|1

:CALCulate:DDEMod:NORMalize:?

368

Digital Demod

Meas Setup

Example

Key Path

Mode

Preset

State Saved

CALC:DDEM:NORM ON

CALC:DDEM:NORM?

Meas Setup, Advanced Dig Demod

VSA

ON

Saved in instrument state.

APSK R2 / R1

This allows you to specify the expected ratio between the two inner rings for APSK modulation formats.

APSK R2 / R1 determines the Ring 2 to Ring 1 ratio for APSK format measurements.

The ring ratio is the ratio of the magnitude of symbol states on a ring (R2) to the magnitude of symbol states on the inner ring (R1). R2 / R1 is a valid parameter for both 16 APSK and 32 APSK format measurements.

Remote Command

Example

Key Path

Mode

Notes

Preset

State Saved

Min

Max

[:SENSe]:DDEMod:APSK:R2Ratio <real>

[:SENSe]:DDEMod:APSK:R2Ratio?

DDEM:APSK:R2R 3

DDEM:APSK:R2R?

Meas Setup, Advanced Dig Demod

VSA

Affects the position of the ideal state indicators on constellation diagrams

2.84

Saved in instrument state.

1.25

8

APSK R3 / R1

This allows you to specify the expected ratio between the outer and inner rings for APSK32 modulation formats.

APSK R3 / R1 determines the Ring 3 to Ring 1 ratio for APSK format measurements.

The ring ratio is the ratio of the magnitude of symbol states on a ring (R3) to the magnitude of symbol states on the inner ring (R1). R3 / R1 is a valid parameter only for 32 APSK format measurements.

Remote Command

[:SENSe]:DDEMod:APSK:R3Ratio <real>

[:SENSe]:DDEMod:APSK:R3Ratio?

369

Digital Demod

Meas Setup

Example

Key Path

Mode

Notes

Preset

State Saved

Min

Max

DDEM:APSK:R3R 5

DDEM:APSK:R3R?

Meas Setup, Advanced Dig Demod

VSA

Affects the position of the ideal state indicators on constellation diagrams

5.27

Saved in instrument state.

1.5625

10

Low SNR Enhancement

Enhances the ability of the demodulator to lock on to signals with low SNR. This process reduces the frequency lock range and enables additional filtering. This filtering allows the demodulator to lock in the presence of more. To compensate for the smaller frequency lock range, the frequency estimate is tracked from measurement to measurement. An exponential average is used and the output of this average becomes the starting point for the next frequency estimate for the next measurement.

This enhancement is only available for the following modulation formats:

• All QAM and DVB QAM formats

• BPSK, QPSK, and 8PSK

• All APSK (amplitude/phase shift keying) and DVB APSK formats

• EDGE

Remote Command

Example

Key Path

Mode

Notes

Preset

State Saved

[:SENSe]:DDEMod:LSNR OFF|ON|0|1

[:SENSe]:DDEMod:LSNR?

DDEM:LSNR OFF

DDEM:LSNR?

Meas Setup, Advanced Dig Demod

VSA

Softkey is greyed out for formats not supported.

OFF

Saved in instrument state.

Adaptive Equalizer Setup

This key displays a menu that allows you to select Adaptive Equalizer parameters for the current measurement. Adaptive equalization removes linear errors from modulated signals by dynamically creating and applying a FIR (feed-forward) compensating filter. Linear errors can come from filters in a

370

Digital Demod

Meas Setup

transmitter or receiver's IF, or from the presence of multiple paths in the transmission path, such as reflections in a cable system. These types of problems appear as group-delay distortion, frequency-response errors (tilt, ripple), and reflections or multipath distortion.

Equalization allows measurement of some impaired channels and can be used to isolate linear from nonlinear error mechanisms. Equalization does not require symbol lock or prior knowledge of the signal

(such as a training sequence) and is compatible with recorded data.

By default, the equalization filter has a unit impulse response which yields a flat frequency response

(only one tap in the filter has a non-zero value and data simply passes through the filter). The position of the unit impulse is a function of the filter length and is positioned to provide the most optimum efficiency for most situations. The position cannot be adjusted.

The equalization filter has a unit impulse response when you: first run the application reset the equalizer filter change points/symbol. change the measured or reference filter change the symbol rate change the clock delay adjustment change the equalizer filter length preset the application

Aside from the above conditions, the application uses the last computed coefficients when you enable equalization. For example, if you used equalization in a previous measurement, the application uses the coefficients from the previous measurement unless you select reset the equalization filter or change

[points/symbol]. Therefore, it is good practice to reset the equalization filter to initialize the filter coefficients before you start a measurement.

Key Path

Mode

Meas Setup

VSA

Filter

Turns adaptive equalization filter on or off. Adaptive equalization uses the measured signal to determine the coefficients of the equalization filter.

When equalization is, the equalization filter has a unit impulse response. The length of the filter determines the position of the unit impulse response in the filter. The impulse is located in the center of the filter for short filter lengths. As the filter length increases, the impulse moves, proportionally, towards the start of the filter to handle channels with large delay-spread.

If Adaptive is set to run, the analyzer uses the results of the current measurement to update the filter coefficients for the next measurement. The analyzer chooses coefficients that produce a modulation quality metric that is less impacted by the presence of linear distortion.

If Adaptive is set to hold, the analyzer does not update the filter coefficients. Instead, the analyzer uses

371

Digital Demod

Meas Setup

the last updated coefficients before selecting hold.

You can select run or hold at any time to continue or stop updating filter coefficients.

Note that the analyzer does not redefine the equalization filter to have a unit impulse response when you select run or when you turn the equalization filter off and then on instead, the analyzer uses the last updated filter coefficients.

Equalization is applied to time-domain data. To see where equalization is applied, see the block diagram for [Digital Demodulation] in this block diagram, equalization is applied in the time-domain corrections block.

For best results, make sure you select a frequency span that contains all energy of your signal. If significant energy from your signal falls outside of the displayed frequency span, equalization will not work on your signal.

You can define the length of the equalization filter (in symbols) and set the convergence (convergence determines the size of the steps used to reshape the equalization filter). For additional details about these parameters, see online help for convergence and filter length.

You can use equalization with pulse search and sync search. Note, however, that the adaptive equalizer does not update the filter coefficients when a "Pulse Not Found" condition exists. In other words, if pulse search is on and the measurement results in a "pulse not found" message, the results of that measurement will not affect the equalization filter coefficients. The same is true if sync search is on and the measurement results in a "SYNC NOT FOUND" message.

The following parameters affect measurement speed when using adaptive equalization: result length filter length (for the equalization filter) points/symbol

Remote Command

Example

Key Path

Mode

Preset

State Saved

[:SENSe]:DDEMod:EQUalization:STATe OFF|ON|0|1

[:SENSe]:DDEMod:EQUalization:STATe?

DDEM:EQU:STAT ON

DDEM:EQU:STAT?

Meas Setup, Adaptive Equalizer Setup

VSA

OFF

Saved in instrument state.

Filter Length

Sets the length (in symbols) of the for the analyzer's equalization filter.

In general, the best filter length is the smallest that meets your measurement requirements. For measurements at the transmitter, the filter length may only need to be a few symbols in length. Longer filter lengths may be needed to measure multi-path environments.

372

Digital Demod

Meas Setup

The filter length also determines the placement of the impulse response in the equalization filter. For longer filter lengths, the analyzer puts the initial, unit impulse response closer to the beginning of the time record to accommodate multi-path measurements, as follows:

Filter Length

(symbol)

3 to 31

31 to 75

75 to 99

Unit Impulse Response Position

(symbol)

(length – 1)/2

15

(length)/5

For example, if the filter length is 11, the unit impulse response is positioned at symbol 5. If the filter length is 35, the unit impulse response is positioned at symbol 15.

Remote Command

Example

Key Path

Mode

Notes

Preset

State Saved

Min

Max

[:SENSe]:DDEMod:EQUalization:FLENgth <integer>

[:SENSe]:DDEMod:EQUalization:FLENgth?

DDEM:EQU:FLEN 31

DDEM:EQU:FLEN?

Meas Setup, Adaptive Equalizer Setup

VSA

Must be an odd number. If an even number is entered, it is rounded up to the next odd.

21

Saved in instrument state.

3

99

Convergence

Sets the adaptive filter convergence factor higher to converge faster. Note that too high a value can cause the filter to not converge. Set convergence factor smaller for better accuracy.

Remote Command

Example

Key Path

Mode

[:SENSe]:DDEMod:EQUalization:CONVergence <real>

[:SENSe]:DDEMod:EQUalization:CONVergence?

DDEM:EQU:CONV 2

DDEM:EQU:CONV?

Meas Setup, Adaptive Equalizer Setup

VSA

373

Digital Demod

Meas Setup

Preset

State Saved

Min

Max

1

Saved in instrument state.

0

10000000

Hold

Turns the filter coefficient updates on or off. Normally the adaptation algorithm updates the filter coefficients after each scan. When Hold is on, the coefficients of the equalization filter are frozen, that is., the adaptive filter becomes fixed. When you turn Hold off again, the coefficients are again allowed to adapt, starting from where they currently are.

Remote Command

Example

Key Path

Mode

Preset

State Saved

[:SENSe]:DDEMod:EQUalization:HOLD OFF|ON|0|1

[:SENSe]:DDEMod:EQUalization:HOLD?

DDEM:EQU:HOLD ON

DDEM:EQU:HOLD?

Meas Setup, Adaptive Equalizer Setup

VSA

OFF

Saved in instrument state.

Reset Filter Coefficients

This function resets the adaptive filter coefficients to 1.

Remote Command

Example

Key Path

Mode

[:SENSe]:DDEMod:EQUalization:RESet

DDEM:EQU:RES

Meas Setup, Adaptive Equalizer Setup

VSA

Preset to Standard

This key allows you to preset Digital Demod Setup parameters and Span to measure a wide variety of standard digital communications formats. Preset to Standard does not constrain you from afterward making any possible adjustment to measurement parameters. The following standard presets are available:

• Cellular: IS–95 Base and Mobile, GSM, EDGE, CDPD, NADC, PDC, PHP, 3GPP (W-CDMA)

• Wireless Networking: 802.11b, HIPERLAN/1 (HBR and LBR), Bluetooth, ZigBee 868, 915, and

2450

• Digital Video: DTV8, DTV16, DVB16, DVB32, DVB64, DVB128, DVB256, DVB 16APSK with

374

Digital Demod

Meas Setup

code rates 2/3 to 9/10, DVB 32 APSK with code rates 3/4 to 9/10.

• Other: APCO 25, DECT, TETRA, VDL Mode 3

Remote Command

Example

Key Path

Mode

[:SENSe]:DDEMod:STANdard:PRESet

CDMABTS|CDMAMS|CDPD|EDGE|GSM|NADC|PDC|PHS|WCDMA|BLUETOO

TH|HIPERLANHBR|HIPERLANLBR|WLAN11B|ZIGBEE2450|ZIGBEE868

|ZIGBEE915|DTV8|DTV16|DVB16|DVB32|DVB64|DVB128|DVB256|D

VB16APSK23|DVB16APSK34|DVB16APSK45|DVB16APSK56|DVB16APS

K89|DVB16APSK910|DVB32APSK34|DVB32APSK45|DVB32APSK56|DV

B32APSK89|DVB32APSK910|APCO|DECT|TETRA|VDL3

DDEM:STAN:PRES CDMABTS

Meas Setup

VSA

Meas Preset

This key performs the same function as Meas Setup, Preset to Standard, W-CDMA. For more information, see the section under the Preset key in the Utility section.

Key Path

Mode

Meas Setup

VSA

375

Digital Demod

Mode

Mode

See

“Mode” on page 213 in the section "Common Measurement Functions" for a description of this

function.

376

Digital Demod

Mode Setup

Mode Setup

See “Mode Setup” on page 229 in the section "Common Measurement Functions" for a description of

this function.

377

Digital Demod

Peak Search

Peak Search

See

“Peak Search” on page 406

in the section "Common Measurement Functions 2" for a description of this function.

378

Digital Demod

Single

Single

See “Single (Single Measurement/Sweep)” on page 263

in the section "Common Measurement

Functions" for a description of this function.

379

Digital Demod

Source

Source

See

“Source” on page 263 in the section "Common Measurement Functions" for a description of this

function.

380

Digital Demod

SPAN X Scale

SPAN X Scale

See “SPAN X Scale” on page 416

in the section "Common Measurement Functions 2" for a description of this function.

381

Digital Demod

Sweep/Control

Sweep/Control

See

“Sweep/Control” on page 424

in the section "Common Measurement Functions 2" for a description of this function.

382

Digital Demod

Trace/Detector

Trace/Detector

This section details the trace results accessible via the data key and also via SCPI, many of which are available in other measurements as well.

See “Trace/Detector” on page 428 in the section "Common Measurement Functions" for a description of

this function.

Select Trace

See “Select Trace” on page 428 in the section "Common Measurement Functions 2" for a description of

this function.

Data

This key provides a menu of trace data choices for the selected trace. For the SCPI command, and other details, see Trace/Detector,

“Data” on page 430

.

The following trace data results are available for all modulation formats but FSK:

Trace data soft key name

SCPI string form

Spectrum "Spectrum1"

Description

Inst Spectrum

Search Time

Time

Raw Main Time

Inst IQ Meas

Spectrum

IQ Ref Time

IQ Ref Spectrum

"Inst Spectrum1"

"Search Time1"

"Time1"

"Raw Main Time1"

IQ Meas Time

IQ Meas Spectrum

"IQ Meas Time1"

"IQ Meas Spec1"

"Inst IQ Meas Spec1"

"IQ Ref Time1"

"IQ Ref Spec1"

Averaged result of successive Inst Spectrum results (If RMS or Max averaging is on). Otherwise, FFT of current windowed Main (or Gate)

Time

Instantaneous Spectrum is the FFT of the current windowed Main (or

Gate) Time. It is instantaneous in the sense that it isn't rms averaged, but it may included time-averaged data.

Entire time data (after corrections) that will be searched (when Burst search or Sync search is on).

Time data used as input to demodulation. If Burst or Sync search is on, this is the time data that fulfills the search criteria.

Time record as it comes from the hardware, before software resampling or corrections

Demodulated Time Trace

Averaged result of successive Inst IQ Meas Spectrum if RMS or Max averaging is on. Otherwise, same as Inst IQ Meas Spectrum

Instantaneous (not averaged) FFT of current IQ Meas Time

Reconstructed ideal time waveform to compare IQ Meas Time against

Averaged result of successive Inst IQ Ref Spectrum if RMS or Max averaging is on. Otherwise, same as Inst IQ Ref Spectrum

383

Digital Demod

Trace/Detector

Trace data soft key name

SCPI string form

Inst IQ Ref

Spectrum

Error Vector Time

"Inst IQ Ref Spec1"

"Error Vector Time1"

Description

Instantaneous (not averaged) FFT of IQ Ref Time

Error Vector

Spectrum

Inst Error Vector

Spectrum

IQ Mag Error

"Error Vector Spec1"

"Inst Err Vect Spec1"

Vector difference between IQ Meas Time and IQ Ref Time at each point in time.

Averaged result of Inst Error Vector Spectrum if RMS or Max averaging is on. Otherwise, same as Inst Error Vector Spectrum.

FFT of Error Vector Time

IQ Phase Error

"IQ Mag Error1"

"IQ Phase Error1"

Equalizer Impulse

Response

Channel Frequency

Response

Symbols/Errors

"Eq Impulse

Response1"

"Ch Frequency

Response1"

"Syms/Errs1"

Difference in length of the IQ Meas Time vector and IQ Ref Time vector at each point in time. Expressed as a % of Ref Time length.

Difference in phase of the IQ Meas Time vector and IQ Ref Time vector at each point in time.

Impulse response of the adaptive equalizer (no data is available if equalizer is off)

FFT of Equalizer Freq Response

Shows table of error statistics and a listing of symbol data. Only symbol results can be obtained using CALC:DDEM:DATA (see Table

Data for how to retrieve tabular results).

Blank Trace No Data "No Data"

If the modulation format is FSK, then the following replace the IQ measurement and reference time and spectrum data, and error vector magnitude data:

FSK Meas Time

FSK Meas Spectrum

Inst FSK Meas

Spectrum

"FSK Meas Time1"

"FSK Meas Spec1"

Demodulated Time Trace

Averaged result of successive Inst FSK Meas Spectrum if RMS or

Max averaging is on. Otherwise, same as Inst FSK Meas Spectrum

"Inst FSK Meas Spec1" Instantaneous (not averaged) FFT of FSK Meas Time

FSK Ref Time "FSK Ref Time1"

FSK Ref Spectrum "FSK Ref Spec1"

"Inst FSK Ref Spec1"

Reconstructed ideal time waveform to compare FSK Meas Time against

Averaged result of successive Inst FSK Ref Spectrum if RMS or Max averaging is on. Otherwise, same as Inst FSK Ref Spectrum

Instantaneous (not averaged) FFT of FSK Ref Time Inst FSK Ref

Spectrum

FSK Error Time "FSK Error Time1" Difference between FSK Meas Time and FSK Ref Time at each point in time.

384

Digital Demod

Trace/Detector

FSK Error Spectrum "FSK Error Spec1" Averaged result of successive Inst FSK Error Spectrum if RMS or

Max averaging is on. Otherwise, same as Inst FSK Error Spectrum.

Instantaneous (not averaged) FFT of FSK Error Spectrum Inst FSK Error

Spectrum

Carrier Mag Error

"Inst FSK Err Spec1"

"Carrier Mag Error1" Amplitude error of carrier, relative to average amplitude.

For more information see: Trace/Detector, “Data” on page 430

Key Path

Mode

Trace/Detector

VSA

Pre-demod

Displays a menu that allows you to select a channel for assignment of trace Data.

Key Path

Mode

Trace/Detector, Data,

VSA

Spectrum

See Trace/Detector, Data,

“Spectrum” on page 431

in the section "Common Measurement Functions 2" for more information.

Inst Spectrum

See Trace/Detector, Data, “Inst Spectrum” on page 433 in the section "Common Measurement Functions

2" for more information.

Search Time

Entire time data (after corrections) that will be searched (when Burst search or Sync search is on).

Key Path

Mode

Trace/Detector, Data, Pre-demod

VSA

Time

Time data used as input to demodulation. If Burst or Sync search is on, this is the time data that fulfills the search criteria.

Key Path

Mode

Trace/Detector, Data, Pre-demod

VSA

Raw Main Time

See Trace/Detector, Data,

“Raw Main Time” on page 436

in the section "Common Measurement

Functions 2" for more information.

385

Digital Demod

Trace/Detector

Demod

Key Path

Mode

IQ Meas Time

Demodulated Time Trace.

Trace/Detector, Data,

VSA

Key Path

Mode

Trace/Detector, Data, Demod

VSA

IQ Meas Spectrum

Averaged result of successive Inst IQ Meas Spectrum if RMS or Max averaging is on. Otherwise, same as Inst IQ Meas Spectrum.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

Inst IQ Meas Spectrum

Instantaneous (not averaged) FFT of current IQ Meas Time.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

IQ Ref Time

Reconstructed ideal time waveform to compare IQ Meas Time against.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

IQ Ref Spectrum

Averaged result of successive Inst IQ Ref Spectrum if RMS or Max averaging is on. Otherwise, same as

Inst IQ Ref Spectrum.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

386

Digital Demod

Trace/Detector

Inst IQ Ref Spectrum

Instantaneous (not averaged) FFT of IQ Ref Time.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

FSK Meas Time

Demodulated Time Trace.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

FSK Meas Spectrum

Averaged result of successive Inst FSK Meas Spectrum if RMS or Max averaging is on. Otherwise, same as Inst FSK Meas Spectrum.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

Inst FSK Meas Spectrum

Instantaneous (not averaged) FFT of FSK Meas Time.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

FSK Ref Time

Reconstructed ideal time waveform to compare FSK Meas Time against.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

FSK Ref Spectrum

Averaged result of successive Inst FSK Ref Spectrum if RMS or Max averaging is on. Otherwise, same as Inst FSK Ref Spectrum.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

387

Digital Demod

Trace/Detector

Inst FSK Ref Spectrum

Instantaneous (not averaged) FFT of FSK Ref Time.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

DemodError

Key Path

Mode

Trace/Detector, Data,

VSA

Error Vector Time

Vector difference between IQ Meas Time and IQ Ref Time at each point in time.

Key Path

Mode

Trace/Detector, Data, DemodError

VSA

Error Vector Spectrum

Averaged result of Inst Error Vector Spectrum if RMS or Max averaging is on. Otherwise, same as Inst

Error Vector Spectrum.

Key Path

Mode

Trace/Detector, Data, DemodError

VSA

Inst Error Vector Spectrum

FFT of Error Vector Time.

Key Path

Mode

Trace/Detector, Data, DemodError

VSA

IQ Mag Error

Difference in length of the IQ Meas Time vector and IQ Ref Time vector at each point in time. Expressed as a % of Ref Time length.

Key Path

Mode

Trace/Detector, Data, DemodError

VSA

388

Digital Demod

Trace/Detector

IQ Phase Error

Difference in phase of the IQ Meas Time vector and IQ Ref Time vector at each point in time.

Key Path

Mode

Trace/Detector, Data, DemodError

VSA

FSK Error Time

Difference between FSK Meas Time and FSK Ref Time at each point in time.

Key Path

Mode

Trace/Detector, Data, DemodError

VSA

FSK Error Spectrum

Averaged result of successive Inst FSK Error Spectrum if RMS or Max averaging is on. Otherwise, same as Inst FSK Error Spectrum.

Key Path

Mode

Trace/Detector, Data, DemodError

VSA

Inst FSK Error Spectrum

Instantaneous (not averaged) FFT of FSK Error Spectrum.

Key Path

Mode

Trace/Detector, Data, DemodError

VSA

Carrier Mag Error

mplitude error of carrier, relative to average amplitude.

Key Path

Mode

Trace/Detector, Data, DemodError

VSA

Table data

One table is available in the Digital Demod measurement. It is displayed when you choose

Symbols/Errors as trace data. The available tabular data changes depending on the modulation format chosen (see Meas Setup, Digital Demod, Format). These values may be obtained using the

CALC:DDEM:DATA:TABL commands (see

Result name

EVM (rms)

Available in Demod

Format:

Displayed Unit

All but FSK %rms

Remote Name

EvmRms

Remote Unit

%rms

389

Digital Demod

Trace/Detector

Result name

EVM (peak)

EVM (peak) symbol number

Offset EVM (rms)

Offset EVM (peak)

Offset EVM (peak) symbol number

FSK error (rms)

FSK error(peak)

FSK error (peak) symbol number

Mag error (rms)

Mag error (peak)

Mag error (peak) symbol number

Phase error (rms)

Phase error (peak)

Phase error (peak) symbol number

Frequency Error

Carrier Offset

SNR(MER)

FSK Deviation

Pilot Level

TimeOffset

IQ Offset

Amplitude Droop

Rho

Quadrature Error

Available in Demod

Format:

Displayed Unit

All but FSK

All but FSK

All

All

All

OQPSK

OQPSK

OQPSK

FSK

FSK

FSK

All but FSK

All but FSK

All but FSK

%pk

%rms

%pk

%rms

%pk

%rms

%pk deg deg

All but FSK

FSK

Hz

Hz

QPSK, QAM, APSK,

VSB dB

FSK

8VSB

Hz dB

APSK (triggered) s

All but FSK or VSB dB dB/sym MSK2, PSK (except

QPSK, OQPSK)

QPSK, OQPSK

All but BPSK, VSB,

FSK deg

Remote Name Remote Unit

PhaseErrRms

PhaseErrPeak

PhaseErrPeakSym

FreqErr

FskCarrOffs

SigToNoise

FskDev

PilotLevel

TimeOffset

IqOffset

AmpDroop

Rho

QuadErr

EvmPeak

EvmPeakSym

OffsetEvmRms

OffsetEvmPeak

OffsetEvmPeakSym

%pk

%rms

%pk

%rms

%pk

FskErrRms

FskErrPeak

FskErrPeakSym

MagErrRms

MagErrPeak

MagErrPeakSym

%rms

%pk deg deg deg

Hz

Hz dB

Hz dB s dB dB/sym

390

Result name Available in Demod

Format:

Displayed Unit

IQ Gain Imbalance All but BPSK, VSB,

FSK

Ring2 to Ring1 Ratio APSK

Ring3 to Ring1 Ratio APSK32

Peak EVM (rms)

95% EVM

EDGE

EDGE dB

%

%

Digital Demod

Trace/Detector

Remote Name Remote Unit

IqGainImbalance dB

R2Ratio

R3Ratio

PeakEvmRms

NinetyFivePercentEv m

%

%

Key Path

Mode

Response

Trace/Detector

VSA

Key Path

Mode

Trace/Detector, Data,

VSA

Equalizer Impulse Response

Impulse response of the adaptive equalizer (no data is available if equalizer is off).

Key Path

Mode

Channel Frequency Response

FFT of Equalizer Freq Response.

Trace/Detector, Data, Response

VSA

Key Path

Mode

Trace/Detector, Data, Response

VSA

Format

This key provides a menu of Trace Formats from which to choose for the selected trace.

See Trace/Detector,

“Format” on page 443

in the section "Common Measurement Functions 2" for more information.

Digital Demod Trace Setup

This key provides a menu that allows you to set Digital Demod parameters for the selected trace.

391

Digital Demod

Trace/Detector

See Trace/Detector,

“Digital Demod Trace Setup” on page 444

in the section "Common Measurement

Functions 2" for more information.

Copy to Data Register

This key provides a menu of Data Registers numbered from 1 to 6 into which to copy the selected trace.

See Trace/Detector,

“Copy to Data Register” on page 450 in the section "Common Measurement

Functions 2" for more information.

Phase/Delay Properties

This key provides a menu that allows you to set Phase and Delay properties for the selected trace.

See Trace/Detector,

“Phase/Delay Properties” on page 451

in the section "Common Measurement

Functions 2" for more information.

ACP Setup

This key provides a menu that allows you to set ACP setup parameters for the selected trace.

See Trace/Detector,

“ACP Setup” on page 453 in the section "Common Measurement Functions 2" for

more information.

OBW Setup

This key provides a menu that allows you to set OBW setup parameters for the selected trace.

See Trace/Detector,

“OBW Setup (Occupied Bandwidth)” on page 463 in the section "Common

Measurement Functions 2" for more information.

Trace Indicator Info

Pressing this key invokes the display of the currently selected trace indicator in the MSG bar at the bottom of the screen.

See Trace/Detector, “Trace Indicator Info” on page 469 in the section "Common Measurement Functions

2" for more information.

392

Digital Demod

Trigger

Trigger

Triggering is used to determine when a measurement should start taking data. There are several available trigger sources. For each trigger source, there are associated setup parameters. Typically, a trigger event is generated when a signal (or a characteristic of the signal) crosses a defined trigger level (or threshold) on a rising or falling slope. The measurement begins at a specified time delay from the trigger point. The delay may be negative, allowing pretrigger data to be taken. Each trigger source has associated its own trigger level, slope, and delay settings.

See “Trigger” on page 470 in the section "Common Measurement Functions 2" for a description of this

function.

393

Digital Demod

View/Display (View Presets)

View/Display (View Presets)

The View/Display key provides access to a menu that enable you to select display parameters for the current measurement.

For more information on other View/Display functions (Display and Layout) see Analyzer Setup,

View/Display

View Presets affect the trace layout, trace data assignment, scaling and formatting but do not affect hardware measurement setup.

Remote Command

Example

Key Path

Mode

:DISPlay:DDEMod:VIEW:PRESet QUAD

DISP:DDEM:VIEW:PRES QUAD

View/Display

VSA

Display

See

“Display” on page 349

in the section "Common Measurement Functions" for a description of this function.

Layout

See

“Layout” on page 479 in the section "Common Measurement Functions 2" for a description of this

function.

Preset View: Demod Quad

This preset displays a quad layout with the IQ Meas Time in trace 1, Spectrum in trace 2, Error Vector

Time in trace 3, and the Symbol/Error table in trace 4.

Key Path

Mode

View/Display

VSA

394

Digital Demod

View/Display (View Presets)

395

Digital Demod

View/Display (View Presets)

396

10

WLAN DSSS

This key selects the VSA WLAN DSSS measurement.

This topic contains the following sections:

“Remote Command Results for VSA WLAN DSS Measurement” on page 397

“Front Panel Results” on page 400

Remote Command Results for VSA WLAN DSS Measurement

All of the scalar results for this measurement are contained one table and hence one equivalent subopcode for the remote results. The remote user can use this equivalence by either visually inspecting the corresponding summary trace on the display, or by using CALC:DATA queries which programmatically describe the corresponding summary trace.

:CONFigure:W11B

:CONFigure:W11B:NDEFault

:FETCh:W11B[n]?

:INITiate:W11B

:MEASure:W11B[n]?

:READ:W11B[n]?

This standard remote result is also available thru the CALC:DATA<n> set of queries, where <n> is a reference to the trace number. The results assigned to each trace vary depending on which tests are enabled. As an example, with the default trace layout, these results in the Error Summary results are returned by CALC:W11B:DATA4:TABLE?

See Common Functions, Data Queries, CALCulate:DATA for more details.

The following table denotes the VSA WLAN DSSS specific results returned from the

(FETCh|MEASure|READ):W11B commands, and their corresponding CALC:DATA queries, indexed by subopcode.

Note that valid results are only returned if the Symbols/Errors trace is being computed. It must be selected though it is not necessary for it to be shown in the current Layout.

Note that there are more results available that are supported by the VSA Mode. These are documented under Common Functions, Data Queries, MEASure, READ, FETCh.

397

WLAN DSSS

Results table (with subopcodes):

N

Not specified or n=1

Results Returned

Returns 22 comma-separated scalar results, corresponding exactly to the items returned in the Error Summary:

1. EVM (% rms)

2. EVM Peak (% peak)

3. EVM Peak Location (chip)

4. Magnitude Error (%rms)

5. Magnitude Error Peak (%)

6. Magnitude Error Peak Location (chip)

7. Phase Error (deg)

8. Phase Error Peak (deg)

9. Phase Error Peak Location (chip)

10. Frequency Error (Hz)

11. IQ Offset (ratio)

12. IQ Quadrature Error (deg)

13. IQ Gain Imbalance (ratio)

14. Sync Correlation (1.0 = ideal)

15. 802.11b EVM Peak (%)

16. Header Status (see below)

17. Mac Status (see below)

18. Burst Type (see below)

19. Bit Rate (bps)

20. Octets

21. Data Time Length (sec)

22. Symbol Clock Error (ratio)

If the results are not available, NaN (9.91e37) is returned.

In addition to these results, other results are defined in Common Functions, Data Queries, MEASure,

READ, FETCh.

Header Status enumerations:

Enumeration Meaning

0 No SFD found - the Preamble Start Frame Delimiter was not found.

398

WLAN DSSS

Enumeration Meaning

1

2

3

4

5

6

Invalid header bits - some of the bits in the PLCP Header were incorrect, but the analyzer was able to determine the data length and data modulation format.

Header CRC failed - the header looked correct except for the CRC, which did not match the rest of the header data.

Invalid header phase shift - an extra (incorrect) 90 degree phase shift was detected at the start of the header.

Invalid Barker1/short - the detected burst type is 1MB/sec Barker with a short preamble, which is an invalid format.

Unknown - the result length was not long enough to capture all of the header data, so the status can not be determined.

OK - the header appears correct and the CRC passed.

Mac Status enumerations:

1

2

Enumeration Meaning

0 Unknown - the result length was not long enough to capture all of the data, so the MAC status can not be determined.

MAC FCS failed - the MAC frame checksum failed.

OK - the MAC frame checksum passed.

Burst Type enumerations:

NOTE

The Burst Type value encodes whether the preamble is the "long format" or "short format", which affects the output value. If the long preamble format is detected, the values are the first number shown. If a short preamble format is detected,

4096 is added to the "long format" value and shown in (parentheses).

Enumeration Meaning

0 (4096)

1 (4097)

2 (4098)

3 (4099)

4 (4100)

5 (4101)

6 (4102)

1 Mbit/sec Barker spreading (BPSK chips)

2 Mbit/sec Barker spreading (QPSK chips)

5.5 Mbit/sec CCK modulation (QPSK chips)

11 Mbit/sec CCK modulation (QPSK chips)

5.5 Mbit/sec PBCC modulation (QPSK chips)

11 Mbit/sec PBCC modulation (QPSK chips)

22 Mbit/sec PBCC modulation (8PSK chips)

399

WLAN DSSS

Enumeration Meaning

7 (4103)

8 (4104)

9 (4105)

33 Mbit/sec PBCC modulation (8PSK chips at 16.5 MHz)

DSSS-OFDM modulation (OFDM symbols)

Unknown modulation format

Key Path

Mode

Meas

VSA

Front Panel Results

VSA WLAN DSSS results may be displayed in any trace, and the traces viewed in a variety of layouts that show 1, 2, 3, or 4 traces at a time. Each trace may be scaled as desired regardless of measurement settings, or auto-scaled to reflect measurement settings. Data may be formatted in a variety of ways. (For example, you may view the log magnitude of complex data, the real or imaginary part, etc.) You may use

Preset View:Basic or other Preset Views to view frequently used results, or to provide a familiar starting point from which you may customize your own view.

400

WLAN DSSS

AMPTD Y Scale

AMPTD Y Scale

See “AMPTD Y Scale (Amplitude)” on page 361 in the section "Common Measurement Functions 2"

for for a description of this function.

401

WLAN DSSS

Auto Couple

Auto Couple

See

“Auto Couple” on page 33 in the section "Common Measurement Functions" for a description of

this function.

402

WLAN DSSS

BW

BW

See “BW (Bandwidth)” on page 368 in the section "Common Measurement Functions 2" for a

description of this function.

FFT Window

See “FFT Window” on page 371

in the section "Common Measurement Functions 2" for a description of this function.

403

WLAN DSSS

Cont (Continuous)

Cont (Continuous)

See

“Cont (Continuous Measurement/Sweep)” on page 35

in the section "Common Measurement

Functions 1" for a description of this function.

404

WLAN DSSS

FREQ Channel

FREQ Channel

See “FREQ Channel” on page 371

in the section "Common Measurement Functions 2" for a description of this function.

405

WLAN DSSS

Input/Output

Input/Output

See

“Input/Output” on page 43

in the section "Common Measurement Functions " for a description of this function.

406

WLAN DSSS

Marker

Marker

See “Marker” on page 374

in the section "Common Measurement Functions 2" for a description of this function.

407

WLAN DSSS

Marker Function

Marker Function

See

“Marker Function” on page 392

in the section "Common Measurement Functions 2" for a description of this function.

408

WLAN DSSS

Marker To

Marker To

See “Marker -> (Marker To)” on page 390 in the section "Common Measurement Functions 2" for a

description of this function.

409

WLAN DSSS

Meas (Measure)

Meas (Measure)

See

“Meas” on page 155 in the section "Common Measurement Functions " for a description of this

function.

410

WLAN DSSS

Meas Setup

Meas Setup

The common Meas Setup functions are documented in

“Meas Setup” on page 400

in the section

"Common Measurement Functions 2".

Key Path

Mode

Front Panel

VSA

Demod Setup

This key brings up a menu of commonly used demodulation setup parameters.

Key Path

Mode

Meas Setup

VSA

Data Modulation Detect

This key allows you to select whether the data modulation format is automatically detected or manually set to a particular data modulation format.

When set to Auto, the SIGNAL and SERVICE field data within the PLCP Header are used to determine the data modulation format. When set to Manual, it is forced to be the value set by Mod Format.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

[:SENSe]:W11B:DEMod:AUTO OFF|ON|0|1

[:SENSe]:W11B:DEMod:AUTO?

Set to Default by Preset to Standard.

W11B:DEM:AUTO ON

Meas Setup, Demod Setup

VSA

ON

Saved in instrument state.

Mod Format

This key sets the modulation format that is forced to be used when Data Modulation Detect is set to

Manual.

Remote Command

[:SENSe]:W11B:DEMod

DSSS1M|DSSS2M|CCK5M5|CCK11M|PBCC5M5|

PBCC11M|PBCC22M|PBCC33M

[:SENSe]:W11B:DEMod?

411

WLAN DSSS

Meas Setup

Coupling

Example

Key Path

Mode

Preset

State Saved

Range

Set to Default by Preset to Standard.

W11B:DEM DSSS1M

Meas Setup, Demod Setup

VSA

DSSS1M

Saved in instrument state.

1 Mbps DSSS | 2 Mbps DSSS | 5.5 Mbps CCK | 11 Mbps CCK | 5.5 Mbps

PBCC | 11 Mbps PBCC | 22 Mbps PBCC | 33 Mbps PBCC

DSSS Formats

This key brings up a selection of the available DSSS Modulation Formats.

Key Path

Mode

Meas Setup, Demod Setup, Mod Format

VSA

1 Mbps DSSS

Selects 1 Mbps 11 Chip DSSS Barker DBPSK Data Modulation Format.

Key Path

Mode

Meas Setup, Demod Setup, Mod Format, DSSS Formats

VSA

2 Mbps DSSS

Selects 2 Mbps 11 Chip DSSS Barker DQPSK Data Modulation Format.

Key Path

Mode

Meas Setup, Demod Setup, Mod Format, DSSS Formats

VSA

CCK Formats

This key brings up a selection of the available CCK Modulation Formats.

Key Path

Mode

Meas Setup, Demod Setup, Mod Format

VSA

5.5 Mbps CCK

Selects 5.5 Mbps 8 Chip CCK DQPSK Data Modulation Format.

Key Path

Mode

Meas Setup, Demod Setup, Mod Format, CCK Formats

VSA

412

11 Mbps CCK

Selects 11 Mbps 8 Chip CCK DQPSK Data Modulation Format.

Key Path

Mode

Meas Setup, Demod Setup, Mod Format, CCK Formats

VSA

PBCC Formats

This key brings up a selection of the available PBCC Modulation Formats.

Key Path

Mode

Meas Setup, Demod Setup, Mod Format

VSA

5.5 Mbps PBCC

Selects 5.5 Mbps PBCC QPSK Data Modulation Format.

Key Path

Mode

Meas Setup, Demod Setup, Mod Format, PBCC Formats

VSA

11 Mbps PBCC

Selects 11 Mbps PBCC QPSK Data Modulation Format.

Key Path

Mode

Meas Setup, Demod Setup, Mod Format, PBCC Formats

VSA

22 Mbps PBCC

Selects 22 Mbps PBCC 8PSK Data Modulation Format.

Key Path

Mode

Meas Setup, Demod Setup, Mod Format, PBCC Formats

VSA

33 Mbps PBCC

Selects 33 Mbps PBCC 8PSK Data Modulation Format.

Key Path

Mode

Meas Setup, Demod Setup, Mod Format, PBCC Formats

VSA

Reference Filter

Selects the Reference Filter, as well as the companion Measurement Filter.

WLAN DSSS

Meas Setup

413

WLAN DSSS

Meas Setup

The following table shows what Measurement filter is used for each selected Reference Filter.

ReferenceFilter CompanionMeasurement Filter

Rect (none)

Gaussian

Raised Cosine

Rect (none)

Rect (none)

Root Raised Cosine (RRC)

The 802.11b standard specifies that no reference filter should be used when computing EVM. However, a transmitter must use a transmit filter in order to meet the 802.11b spectral mask requirement, so it may be useful to specify a reference filter when computing EVM.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Range

[:SENSe]:W11B:FILTer:REFerence

RECTangular|GAUSsian|RCOSine

[:SENSe]:W11B:FILTer:REFerence?

Set to Default by Preset to Standard.

W11B:FILT:REF RECT

Meas Setup, Demod Setup

VSA

RECT

Saved in instrument state.

Rectangular | Gaussian | Raised Cosine

Rectangular

Selects the Rectangular (no) Reference Filter and the Rectangular (no) Measurement Filter

Key Path

Mode

Meas Setup, Demod Setup, Data Modulation, Reference Filter

VSA

Gaussian

Selects the Gaussian Reference Filter and the Rectangular (no) Measurement Filter

Key Path

Mode

Meas Setup, Demod Setup, Data Modulation, Reference Filter

VSA

Raised Cosine

Selects the Raised Cosine Reference Filter and the Root Raised Cosine (RRC) Measurement Filter

Key Path

Meas Setup, Demod Setup, Data Modulation, Reference Filter

414

WLAN DSSS

Meas Setup

Mode VSA

Alpha / BT

If you specify a Gaussian Reference Filter, you can set the BT (Bandwidth Time Product) for the filter.

If you specify a Raised Cosine Reference Filter, you can set the Alpha for the filter.

BT describes the shape of a Gaussian filter. BT indicates the filter roll-off (or excess bandwidth) of the

Gaussian filter.

Alpha describes the shape of a Nyquist (raised cosine) filter. Alpha is also called the roll-off or the excess bandwidth factor. A higher value for alpha increases the bandwidth that is used in excess of the theoretical minimum.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11B:ALPHa <real>

[:SENSe]:W11B:ALPHa?

Set to Default by Preset to Standard.

W11B:ALPH 0.5

Meas Setup, Demod Setup

VSA

0.5

Saved in instrument state.

.05

100.0 (Gaussian Reference Filter), 1.0 (Raised Cosine Reference Filter)

Preset to Standard

This immediately sets all measurement parameters to their Preset values. This is the same as Meas

Preset except that the display state is left alone.

Remote Command

Example

Key Path

Mode

Notes

[:SENSe]:W11B:STANdard:PRESet I11BGDSSS

W11B:STAN:PRES I11BGDSSS

Meas Setup

VSA

The parameter is required although there is only one choice.

Meas Time

This key brings up a menu consisting of the time length and positioning demodulation setup parameters.

Key Path

Meas Setup

415

WLAN DSSS

Meas Setup

Mode VSA

Search Length

Lets you specify the length of time to acquire the input signal (in seconds). This defines the length of time that the pulse search is performed.

The analyzer searches within the specified search length and demodulates the first complete pulse

(burst). The pulse can occur anywhere within the search length. Pulses beyond the first complete pulse are disregarded and are not demodulated. In order for the pulse search algorithm to detect a pulse edge, the pulse must be at least 15 dB above the noise floor. The resultant pulse is then demodulated.

Even if no pulse is found, the analyzer will attempt to demodulate starting at the beginning of the Search

Length. If a PLCP Preamble Sync pattern is found near the start of the search length, the demodulation will be correct.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11B:SYNC:SLENgth <time>

[:SENSe]:W11B:SYNC:SLENgth?

The minimum is constrained to meet the number of chip requirements of

Result Length or Max Auto Result Length, whichever is active, at the current

Chip Rate.

Set to Default by Preset to Standard.

W11B:SYNC:SLEN 0.001

Meas Setup, Meas Time

VSA

1.0 ms

Saved in instrument state.

.001 ms

33.70373 ms

Result Length

This key allows you to manually set the Result Length or to have it automatically detected.

The result length is the total number of chips included in the acquired and demodulated data and includes the PLCP Preamble and Header.

When Result Length is Automatic, the result length is the number of chips in the current burst, up to the limit of Max Auto Result Length.

Note that when Result Length is Manual, the demodulation is not limited to the number of chips transmitted within the burst as is done when the Result Length is Auto. The analyzer will demodulate the total number of chips specified by Result Length, which if longer than the burst, could lead to

416

WLAN DSSS

Meas Setup

unexpected and incorrect data results.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11B:TIME:RESult:LENGth <integer>

[:SENSe]:W11B:TIME:RESult:LENGth?

[:SENSe]:W11B:TIME:RESult:AUTO OFF|ON|0|1

[:SENSe]:W11B:TIME:RESult:AUTO?

Search Length is increased to include the Result Length, if Result Length is in

Manual mode.

Set to Default by Preset to Standard.

W11B:TIME:RES:LENG 2816

W11B:TIME:RES:AUTO ON

Meas Setup, Meas Time

VSA

2816 chips

ON

Saved in instrument state.

1 chip

65941 chips

Max Auto Result Length

This key has an effect only when Result Length is set to Auto. In this case, the effective result length, the total number of chips included in the demodulated data, is automatically determined by comparing the input signal's chip length to the Max Auto Result Length value and using the smaller number.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11B:TIME:RESult:MAX <integer>

[:SENSe]:W11B:TIME:RESult:MAX ?

Search Length is increased to include the Max Result Length, if Result Length is in Auto mode.

Set to Default by Preset to Standard.

W11B:TIME:RESult:MAX 2816

Meas Setup, Meas Time

VSA

2816 chips

Saved in instrument state.

1 chip

370741 chips

417

WLAN DSSS

Meas Setup

Meas Offset

Measurement Offset determines the start position, in chips, of the Meas Interval with respect to the first

Chip of the PLCP Preamble. Measurement Interval is entered in chips, which is rounded to the nearest whole number of data symbols, as determined from the input signal's data modulation format.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11B:TIME:OFFSet <integer>

[:SENSe]:W11B:TIME:OFFSet?

The maximum is limited to Result Length or Max Auto Result Length, whichever is active, minus Meas Interval.

Set to Default by Preset to Standard.

W11B:TIME:OFFS 22

Meas Setup, Meas Time

VSA

22 chips

Saved in instrument state.

0 chips

See coupling

Meas Interval

Measurement Interval, in conjunction with Meas Offset, lets you specify an interval (segment) of the result length data to be demodulated and analyzed. Measurement Interval is entered in chips, which is rounded to the nearest whole number of data symbols, as determined from the input signal's data modulation format.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

[:SENSe]:W11B:TIME:INTerval <integer>

[:SENSe]:W11B:TIME:INTerval?

The maximum is limited to Result Length or Max Auto Result Length, whichever is active. If necessary, Meas Offset is reduced to keep Meas

Interval plus Meas Offset within Result Length or Max Result Length, whichever is active.

Set to Default by Preset to Standard.

W11B:TIME:INT 2794

Meas Setup, Meas Time

VSA

2794 chips

Saved in instrument state.

1

418

WLAN DSSS

Meas Setup

Max See coupling

Advanced

This key accesses a menu that allows you to select lesser used demodulation parameters for the current measurement. These settings are for advanced users and do not normally require adjustment for most common measurements.

Key Path

Mode

Meas Setup, More

VSA

Chip Rate

This key lets you set the Chip Rate for the analyzer's demodulator. This allows you the freedom to enter nonstandard chip rates for test and analysis purposes.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11B:CRATe <freq>

[:SENSe]:W11B:CRATe?

Span is forced to be at least one tenth of the Chip Rate. So the maximum Chip

Rate is one tenth of the maximum Span. Likewise, the Span will be forced to be within ten times the Chip Rate and so the minimum Span sets the minimum

Chip Rate.

Set to Default by Preset to Standard.

W11B:CRAT 11.0E6

Meas Setup, More, Advanced

VSA

11 MHz

Saved in instrument state.

See Coupling

See Coupling

Clock Adjust

Clock Adjust determines when the analyzer's digital demodulator samples the I/Q trajectory.

The demodulator used in this measurement does not require symbol-clock timing signals to determine the location of chip detection-decision points. Instead, the demodulator uses an algorithm to determine chip locations.

Some digital communications systems contain non-linearities that can bias the digital demodulator's estimation of the chip clock position. You can use clock adjust to compensate for this "offset" and obtain

419

WLAN DSSS

Meas Setup

a lower EVM (Error Vector Magnitude).

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11B:CADJust <real>

[:SENSe]:W11B:CADJust?

Set to Default by Preset to Standard.

W11B:CADJ 0.0

Meas Setup, More, Advanced

VSA

0 chips

Saved in instrument state.

–0.5 chips

0.5 chips

Track Phase

The Track Phase parameter enables tracking of phase drift within the WLAN burst. This may be useful if the carrier frequency drifts significantly during a burst. Circular arcs in the constellation of the IQ

Meas Time trace display is an indicator of phase drift. The Track Phase parameter can be used to remove the phase drift, which will improve the IQ constellation trace and reduce the EVM data results.

Frequency drift can also cause poor equalizer performance, so selecting Track Phase may improve equalizer results.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

[:SENSe]:W11B:TRACk:PHASe OFF|ON|0|1

[:SENSe]:W11B:TRACk:PHASe?

Set to Default by Preset to Standard.

W11B:TRAC:PHAS OFF

Meas Setup, More, Advanced

VSA

OFF

Saved in instrument state.

Equalizer

Equalizer State turns the Equalizer on or off.

The Equalizer numeric entry sets the length of the equalization filter.

In general, the best Equalizer filter length is the smallest that meets your measurement requirements. For measurements at the transmitter, the filter length may only need to be a few chips in length. Longer filter lengths may be needed to measure multi-path environments.

The 802.11b specification does not allow for equalization prior to computing EVM. This means that any

420

WLAN DSSS

Meas Setup

linear distortion, such as group delay distortion in the IF, will increase EVM. When EVM is high the equalizer can be used as a diagnostic tool. If use of the equalizer significantly improves the EVM result, then the channel frequency response should be examined for flatness problems (i.e. group delay distortion). If it does not, then the problem is more likely related to noise, non-linear distortion, or spurious error.

Remote Command

Coupling

Restriction and Notes

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11B:EQUalization:FLENgth <integer>

[:SENSe]:W11B:EQUalization:FLENgth?

[:SENSe]:W11B:EQUalization[:STATe] OFF|ON|0|1

[:SENSe]:W11B:EQUalization[:STATe]?

The maximum Equalizer length is also limited to 2 chips less than the result length.

Set to Default by Preset to Standard.

Only odd values are valid. Even value entries are rounded down.

W11B:EQU:FLEN 21

W11B:EQU OFF

Meas Setup, More, Advanced

VSA

21

OFF

Saved in instrument state.

3

99

Descramble

Descramble Mode lets you enable/disable the WLAN 802.11b/g descrambler. With this feature you can separately view the descrambled or raw data bits for the preamble, header or payload data fields.

Remote Command

Coupling

Example

Key Path

Mode

Preset

Range

[:SENSe]:W11B:DESCramble ALL|NONE|PREamble|PHEader

[:SENSe]:W11B:DESCramble?

Set to Default by Preset to Standard.

W11B:DESC ALL

Meas Setup, More, Advanced

VSA

ALL

All | None | Preamble Only | Preamble & Header Only

421

WLAN DSSS

Meas Setup

All

The WLAN Descrambler is ON, the Preamble, Header, and Payload data bits are descrambled.

Key Path

Mode

Meas Setup, More, Advanced, Descramble

VSA

None

The WLAN Descrambler is OFF, the Preamble, Header, and Payload data bits are not descrambled.

Key Path

Mode

Meas Setup, More, Advanced, Descramble

VSA

Preamble Only

Only the Preamble data bits are descrambled. The Header data bits and Payload data bits are not descrambled.

Key Path

Mode

Meas Setup, More, Advanced, Descramble

VSA

Preamble & Header Only

Only the Preamble and Header data bits are descrambled. The Payload data bits are not descrambled.

Key Path

Mode

Meas Setup, More, Advanced, Descramble

VSA

IQ Normalize

When set to On, the IQ trace data results (including IQ Meas Time, IQ Ref Time, IQ Mag Err, IQ Phase

Err, and Error Vector Time and corresponding Spectrum results) are normalized.

When set to Off, the trace data results are not normalized.

When normalization is On, the analyzer normalizes or scales the demodulated trace data results, in magnitude, to a nominal value of 1, relative to the outermost states of the ideal (IQ Reference) constellation diagram.

Note that regardless of this setting, the Error Summary trace data results are always normalized.

Remote Command

Coupling

Example

Key Path

:CALCulate:W11B:NORMalize OFF|ON|0|1

:CALCulate:W11B:NORMalize?

Set to Default by Preset to Standard.

CALC:W11B:NORM ON

Meas Setup, More, Advanced

422

WLAN DSSS

Meas Setup

Mode

Preset

State Saved

VSA

ON

Saved in instrument state.

Meas Preset

This immediately sets all measurement parameters to their Preset values. This presets the display state in the same way as Preset View: Basic.

For more information, see the section under the Preset key in the Utility section.

Key Path

Mode

Meas Setup, More

VSA

423

WLAN DSSS

Mode

Mode

See

“Mode” on page 213 in the section "Common Measurement Functions" for a description of this

function.

424

WLAN DSSS

Mode Setup

Mode Setup

See “Mode Setup” on page 229 in the section "Common Measurement Functions" for a description of

this function.

425

WLAN DSSS

Peak Search

Peak Search

See

“Peak Search” on page 406

in the section "Common Measurement Functions 2" for a description of this function.

426

WLAN DSSS

Single

Single

See “Single (Single Measurement/Sweep)” on page 263

in the section "Common Measurement

Functions" for a description of this function.

427

WLAN DSSS

Source

Source

See

“Source” on page 263 in the section "Common Measurement Functions" for a description of this

function.

428

WLAN DSSS

SPAN X Scale

SPAN X Scale

See “SPAN X Scale” on page 416

in the section "Common Measurement Functions 2" for a description of this function.

429

WLAN DSSS

Sweep/Control

Sweep/Control

See

“Sweep/Control” on page 424

in the section "Common Measurement Functions 2" for a description of this function.

430

WLAN DSSS

Trace/Detector

Trace/Detector

See “Trace/Detector” on page 428 in the section "Common Measurement Functions 2" for a description

of this function.

Select Trace

See “Select Trace” on page 428 in the section "Common Measurement Functions 2" for a description of

this function.

Data

This key provides a menu of trace data choices for the selected trace. For the SCPI command, and other details, see Trace/Detector,

“Data” on page 430

.

The following are the SCPI string forms for the Data types that are specific to this measurement, in alphabetical order:

Trace data soft key name

CCDF

CDF

Ch Frequency Response

Eq Impulse Response

Error Vector Spectrum

Error Vector Time

Header Symbols

IQ Mag Error

IQ Meas Spectrum

IQ Meas Time

IQ Phase Error

IQ Ref Spectrum

IQ Ref Time

Inst Error Vector Spectrum

Inst IQ Meas Spectrum

Inst IQ Ref Spectrum

Inst Spectrum

PDF

SCPI string form

"CCDF1"

"CDF1"

"Ch Frequency Response1"

"Eq Impulse Response1"

"Error Vector Spec1"

"Error Vector Time1"

"Header Syms1"

"IQ Mag Error1"

"IQ Meas Spec1"

"IQ Meas Time1"

"IQ Phase Error1"

"IQ Ref Spec1"

“IQ Ref Time1”

"Inst Error Vect Spec1"

"Inst IQ Meas Spec1"

"Inst IQ Ref Spec1"

"Inst Spectrum1"

"PDF1"

431

WLAN DSSS

Trace/Detector

Trace data soft key name

Preamble Symbols

Raw Main Time

Search Time

Spectrum

Symbols/Errors

Time

SCPI string form

"Preamble Syms1"

"Raw Main Time1"

"Search Time1"

"Spectrum1"

"Syms/Errs1"

"Time1"

Pre Demod

This key accesses the Trace Data choices which show pre-demodulation results.

Key Path

Mode

Trace/Detector, Data

VSA

Spectrum

Averaged FFT of the Time waveform. See Trace/Detector, Data, Pre-demod, “Spectrum” on page 431 in

the section "Common Measurement Functions 2" for more information.

Inst Spectrum

The FFT of the time waveform for the current measurement. “Inst” or Instantaneous refers to this result not being averaged like the Spectrum Trace Data result. See Trace/Detector, Data, Pre-demod,

“Inst

Spectrum” on page 433

in the section "Common Measurement Functions 2" for more information.

Search Time

The Search Length long time record acquired for the current measurement.

Key Path

Mode

Trace/Detector, Data, Pre Demod

VSA

Time

The time record before digital demodulation and after the pulse/burst search.

The starting point is the beginning of the burst, offset by the Meas Offset and its length is the Meas

Interval.

The exception to this is that 10% more additional data on both sides is shown when the Measurement

Offset is set to zero and the Measurement Interval is set the same as the Result Length in Manual Result

Length mode. In automatic Result Length mode, this will occur when the Measurement Offset to zero and the Measurement Interval is set to the Max Result Length, but will not go more than 10% beyond the end of the burst.

Key Path

Trace/Detector, Data, Pre Demod

432

WLAN DSSS

Trace/Detector

Mode VSA

Raw Main Time

The raw time record acquired for the current measurement. This data is unprocessed and includes additional points acquired for settling of the filters involved in subsequent processing, such as the

demodulation filtering. See Trace/Detector, Data, Pre-demod, “Raw Main Time” on page 436

in the section "Common Measurement Functions 2" for more information.

Statistical

This key accesses the Trace Data choices which show statistical results.

Key Path

Mode

Trace/Detector, Data

VSA

CCDF

The Complementary, Cumulative Density function (CCDF) for the selected input channel.

The analyzer plots CCDF using units of percent (%) for the y-axis and power (dB) for the x-axis. Power

on the x-axis is relative to the signal average power. See Trace/Detector, Data, Statistical, “CCDF

(Complementary, Cumulative Density Function)” on page 441 in the section "Common Measurement

Functions 2" for more information.

CDF

The Cumulative Density Function (CDF) for the selected input channel. CDF is computed by integrating the CCDF (Probability Density Function). See Trace/Detector, Data, Statistical,

“CDF (Cumulative

Density Function)” on page 442

in the section "Common Measurement Functions 2" for more information.

PDF

The Probability Density Function (PDF) for the selected input channel. PDF indicates the probability that a given level has occurred. See Trace/Detector, Data, Statistical,

“PDF (Probability Density

Function)” on page 442 in the section "Common Measurement Functions 2" for more information.

Demod

This key accesses the Trace Data choices which show general demodulation results.

Key Path

Mode

Trace/Detector, Data

VSA

IQ Meas Time

The measured time data sampled at exactly the chip times with IQ origin offset, system gain normalization and carrier locking applied to the input signal.

Key Path

Trace/Detector, Data, Demod

433

WLAN DSSS

Trace/Detector

Mode VSA

IQ Meas Spectrum

The averaged frequency spectrum of IQ Meas Time. The selected FFT Window is applied and the FFT performed on the result. This is then averaged, if averaging is turned on.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

Inst IQ Meas Spectrum

The instantaneous (non-averaged) frequency spectrum of IQ Meas Time. The selected FFT Window is applied and the FFT performed on the result.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

IQ Ref Time

The ideal representation of the measured input signal (IQ Meas Time).

Key Path

Mode

Trace/Detector, Data, Demod

VSA

IQ Ref Spectrum

The averaged frequency spectrum of IQ Ref Time. The selected FFT Window is applied and the FFT performed on the result. This is then averaged, if averaging is turned on.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

Inst IQ Ref Spectrum

The instantaneous (non-averaged) frequency spectrum of IQ Ref Time. The selected FFT Window is applied and the FFT performed on the result.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

Demod Error

This key accesses the Trace Data choices which show demodulation error related results.

Key Path

Trace/Detector, Data

434

WLAN DSSS

Trace/Detector

Mode VSA

Error Vector Time

The vector (IQ) difference between the IQ Meas Time and IQ Ref Time signals.

Key Path

Mode

Trace/Detector, Data, Demod Error

VSA

Error Vector Spectrum

The averaged frequency spectrum of Error Vector Time. The selected FFT Window is applied and the

FFT performed on the result. This is then averaged, if averaging is turned on.

Key Path

Mode

Trace/Detector, Data, Demod Error

VSA

Inst Error Vector Spectrum

The instantaneous (non-averaged) frequency spectrum of Error Vector Time. The selected FFT Window is applied and the FFT performed on the result.

Key Path

Mode

Trace/Detector, Data, Demod Error

VSA

IQ Mag Error

The difference, in magnitude, between the IQ Meas Time and IQ Ref Time signals.

Key Path

Mode

Trace/Detector, Data, Demod Error

VSA

IQ Phase Error

The phase difference between the IQ Meas Time and IQ Ref Time signals.

Key Path

Mode

Trace/Detector, Data, Demod Error

VSA

Tables

This key accesses the Trace Data choices which are in tabular form, including demodulated symbols tables.

Key Path

Mode

Trace/Detector, Data

VSA

435

WLAN DSSS

Trace/Detector

Symbols/Errors

This display contains an error summary in the upper part, and a listing of demodulated symbols in the lower part. The error summary consists of:

Result name Displayed

Unit

EVM (rms)

EVM (peak)

EVM (peak) location

Mag error (rms)

Mag error (peak)

Mag error (peak) location

Phase error (rms)

Phase error (peak)

Phase error (peak) location

Frequency Error

IQ Offset

Quadrature Error

Gain Imbalance

Sync Correlation

802.11b 1000 chip Peak EVM %

Status chip

Hz dB deg dB

(included in Status above)

Burst Type

Bit Rate

Octets

Data Len

Symbol Clock Err bps sec ppm

%pk chip deg deg

%rms

%pk chip

%rms

Remote Name Remote

Unit

EVM

EVMPeak

EVMPeakLoc

MagErr

MagErrPeak

MagErrPeakLoc

PhaseErr

PhaseErrPeak

PhaseErrPeakLoc chip

FreqErr Hz

IQOffset

IQQuadErr

(none) deg

IQGainImb

SyncCorr

(none)

(none)

80211bEvmPeak %

HdrStat (none)

% chip deg deg

%rms

% chip

%rms

MacStat

BurstType

BitRate

Octets

DataTimeLen

SymClkErr

(none)

(none) bps

(none) sec

(none)

The error summary values may be obtained using the CALC:W11B:DATA:TABL commands.

The demodulated symbols are available as the data values for this trace and so may be accessed using the

CALCulate:W11B:DATA command. See Common Functions, Data Queries, CALCulate:DATA for more details.

Key Path

Trace/Detector, Data, Tables

436

WLAN DSSS

Trace/Detector

Mode VSA

Preamble Symbols

This trace shows the 802.11b PLCP Preamble data bits.

Key Path

Mode

Trace/Detector, Data, Tables

VSA

Header Symbols

This trace shows the 802.11b PLCP Header data bits.

Key Path

Mode

Trace/Detector, Data, Tables

VSA

Response

This key accesses the Trace Data choices which show equalizer response results.

Key Path

Mode

Trace/Detector, Data

VSA

Ch Frequency Response

When the Equalizer is On, this trace shows the frequency response of the channel for which the equalizer is correcting. Ch Frequency Response is computed as the inverse of the equalization filter's frequency response.

Key Path

Mode

Trace/Detector, Data, Response

VSA

EQ Impulse Response

When the Equalizer is On, the EQ Impulse Response trace shows the impulse response computed from the 802.11b or 802.11g preamble.

Key Path

Mode

Trace/Detector, Data, Response

VSA

ACP

This brings up a selection of the ACP result traces. See Trace/Detector, Data,

“ACP Setup” on page 453

in the section "Common Measurement Functions 2" for more information.

437

WLAN DSSS

Trace/Detector

OBW

This brings up a selection of the OBW result traces. See Trace/Detector, Data,

“OBW Setup (Occupied

Bandwidth)” on page 463 in the section "Common Measurement Functions 2" for more information.

Register

This brings up a selection of the Data Registers. See Trace/Detector, Data,

“Register” on page 468

in the section "Common Measurement Functions 2" for more information.

No Data

A blank display is shown.

Format

See

“Format” on page 443 in the section "Common Measurement Functions 2" for a description of this

function.

Digital Demod Trace Setup

See

“Digital Demod Trace Setup” on page 444

in the section "Common Measurement Functions 2" for a description of this function.

Copy to Data Register

See

“Copy to Data Register” on page 450

in the section "Common Measurement Functions 2" for a description of this function.

Phase/Delay Properties

See

“Phase/Delay Properties” on page 451 in the section "Common Measurement Functions 2" for a

description of this function.

ACP Setup

See

“ACP Setup” on page 453 in the section "Common Measurement Functions 2" for a description of

this function.

OBW Setup

See

“OBW Setup (Occupied Bandwidth)” on page 463 in the section "Common Measurement Functions

2" for a description of this function.

Trace Indicator Info

See

“Trace Indicator Info” on page 469

in the section "Common Measurement Functions 2" for a description of this function.

438

WLAN DSSS

Trigger

Trigger

This key is used to set up the triggering for a measurement. See “Trigger” on page 470 in the section

"Common Measurement Functions 2" for a description of this function.

439

WLAN DSSS

View/Display

View/Display

The view setup can be changed by selections from the View/Display menu, including by pressing View

Preset: Basic.

Key Path

Mode

Front-panel key

VSA

Display

See

“Display” on page 349

in the section "Common Measurement Functions" for a description of this function.

Layout

See

“Layout” on page 479 in the section "Common Measurement Functions 2" for a description of this

function.

Preset View

Remote Command

Example

Key Path

Mode

:DISPlay:W11B:VIEW:PRESet BASic|EVM|POWer|DIAGnostic

DISP:W11B:VIEW:PRES BAS

(SCPI only)

VSA

Preset View: Basic

This preset view consists of the following traces in a Grid 2x2 layout:

1. IQ Meas Time in I-Q format

2. Spectrum in Log Mag (dB) format

3. Search Time in Log Mag (dB) format

4. Symbols/Errors

The Preset View: Basic softkey does an immediate action of changing the layout and view to this state.

This Preset View is an action, not a state.

This layout is the layout set by Meas Preset and is good for insuring that the signal is being demodulated correctly, as well as giving visibility to many basic demodulation setup problems.

Remote Command

Example

:DISPlay:W11B:VIEW:PRESet BASic

DISP:W11B:VIEW:PRES BAS

440

Key Path

Mode

View/Display

VSA

WLAN DSSS

View/Display

Preset View: EVM

This preset view consists of the following traces in a Stacked layout:

1. Error Vector Time in Linear Mag format

2. Symbols/Errors

This layout is for more detailed EVM analysis.

The Preset View: EVM softkey does an immediate action of changing the layout and view to this state.

This Preset View is an action, not a state.

Remote Command

Example

Key Path

:DISPlay:W11B:VIEW:PRESet EVM

DISP:W11B:VIEW:PRES EVM

View/Display

441

WLAN DSSS

View/Display

Mode VSA

Preset View: Power

This preset view consists of the following traces in a Grid 2x2 layout:

1. Spectrum in Log Mag (dB) format

2. Time in Log Mag (dB) format

3. Search Time in Log Mag (dB) format

4. CCDF in Log Mag (Linear Unit) format

This layout is oriented towards examining the signal in the Time and Power domain.

The Preset View: Basic softkey does an immediate action of changing the layout and view to this state.

This Preset View is an action, not a state.

Remote Command

Example

:DISPlay:W11B:VIEW:PRESet POWer

DISP:W11B:VIEW:PRES POW

442

Key Path

Mode

View/Display

VSA

WLAN DSSS

View/Display

Preset View: Diagnostic

This preset view consists of the following traces in a Grid 2x2 layout:

1. Error Vector Time in Linear Mag format

2. IQ Meas Time in I-Q format

3. IQ Mag Error in Linear Mag format

4. IQ Phase Error in Wrap Phase format

This layout is useful for diagnosing modulation impairments and errors.

The Preset View: Basic softkey does an immediate action of changing the layout and view to this state.

This Preset View is an action, not a state.

Remote Command

:DISPlay:W11B:VIEW:PRESet DIAGnostic

443

WLAN DSSS

View/Display

Example

Key Path

Mode

DISP:W11B:VIEW:PRES DIAGnostic

View/Display

VSA

444

11

WLAN OFDM

This key selects the VSA WLAN OFDM measurement.

This topic contains the following sections:

“Remote Command Results for WLAN OFDM Measurement” on page 445

“Front Panel Results” on page 446

Remote Command Results for WLAN OFDM Measurement

All of the scalar results for this measurement are contained one table and hence one equivalent subopcode for the remote results. The remote user can use this equivalence by either visually inspecting the corresponding summary trace on the display, or by using CALC:DATA queries which programmatically describe the corresponding summary trace.

:CONFigure:W11A

:CONFigure:W11A:NDEFault

:FETCh:W11A[n]?

:INITiate:W11A

:MEASure:W11A[n]?

:READ:W11A[n]?

This standard remote result is also available thru the CALC:DATA<n> set of queries, where <n> is a reference to the trace number. The results assigned to each trace vary depending on which tests are enabled. As an example, with the default trace layout, these results in the Error Summary results are returned by CALC:W11A:DATA4:TABLE?

See Common Functions, Data Queries,

CALCulate:DATAfor more details.

The following table denotes the VSA WLAN OFDM specific results returned from the

(FETCh|MEASure|READ):W11A commands, and their corresponding CALC:DATA queries, indexed by subopcode.

Note that valid results are only returned if the Symbols/Errors trace is being computed. It must be selected though it is not necessary for it to be shown in the current Layout.

Note that there are more results available that are supported by the VSA Mode. These are documented under Common Functions, Data Queries, MEASure, READ, FETCh.

445

WLAN OFDM

Results table (with subopcodes):

N

Not specified or n=1

Results Returned

Returns 22 comma-separated scalar results, corresponding exactly to the items returned in the Error Summary:

1. EVM (dB)

2. EVM (% rms)

3. Pilot EVM (dB)

4. CPE (%)

5. Frequency Error (Hz)

6. IQ Offset (dB)

7. IQ Quadrature Error (deg)

8. IQ Gain Imbalance (dB)

9. Sync Correlation (1.0 = ideal)

10. Symbol Clock Error (ppm)

11. Modulation Format

12. Octets (bytes)

13. Symbols

14. Code Rate (ratio)

15. Bit Rate (bps)

If the results are not available, NaN (9.91e37) is returned.

In addition to these results, other results are defined in Common Functions, Data Queries, MEASure,

READ, FETCh.

Modulation Format enumerations:

Enumeration Meaning

2

4

0

1

6

Unknown

BPSK

QPSK

16QAM

64QAM

Front Panel Results

VSA WLAN OFDM results may be displayed in any trace, and the traces viewed in a variety of layouts that show 1, 2, 3, or 4 traces at a time. Each trace may be scaled as desired regardless of measurement settings, or auto-scaled to reflect measurement settings. Data may be formatted in a variety of ways. (For

446

WLAN OFDM example, you may view the log magnitude of complex data, the real or imaginary part, etc.) You may use

Preset View: Basic or other Preset Views to view frequently used results, or to provide a familiar starting point from which you may customize your own view.

Key Path

Mode

Meas

VSA

447

WLAN OFDM

AMPTD Y Scale

AMPTD Y Scale

See

“AMPTD Y Scale (Amplitude)” on page 361

in the section "Common Measurement Functions 2" for for a description of this function.

448

Auto Couple

See

“Auto Couple” on page 33

in the section "Common Measurement Functions".

WLAN OFDM

Auto Couple

449

WLAN OFDM

BW

BW

See

“BW (Bandwidth)” on page 368

in the section "Common Measurement Functions 2" for a description of this function.

FFT Window

See “FFT Window” on page 371 in the section "Common Measurement Functions 2" for a description of

this function.

450

WLAN OFDM

Cont (Continuous)

Cont (Continuous)

See “Cont (Continuous Measurement/Sweep)” on page 35 in the section "Common Measurement

Functions 1" for a description of this function.

451

WLAN OFDM

FREQ Channel

FREQ Channel

See

“FREQ Channel” on page 371 in the section "Common Measurement Functions 2" for a description

of this function.

452

WLAN OFDM

Input/Output

Input/Output

See “Input/Output” on page 43 in the section "Common Measurement Functions " for a description of

this function.

453

WLAN OFDM

Marker

Marker

See

“Marker” on page 374 in the section "Common Measurement Functions 2" for a description of this

function.

454

WLAN OFDM

Marker Function

Marker Function

See “Marker Function” on page 392 in the section "Common Measurement Functions 2" for a

description of this function.

455

WLAN OFDM

Marker To

Marker To

See

“Marker -> (Marker To)” on page 390

in the section "Common Measurement Functions 2" for a description of this function.

456

WLAN OFDM

Meas (Measure)

Meas (Measure)

See “Meas” on page 155

in the section "Common Measurement Functions " for a description of this function.

457

WLAN OFDM

Meas Setup

Meas Setup

This key provides a menu allowing you to select measurement parameters for the current measurement.

Key Path

Mode

Front Panel

VSA

Avg Number

See

“Avg Number” on page 400

in the section "Common Measurement Functions 2" for a description of this function.

Averaging applies to a limited set of measurement results in VSA WLAN OFDM. RMS and Max average types apply to the Spectrum and Ch Frequency Response traces. The behavior for these types is the same as in the Vector Analysis Measurement. Averaging of numeric error data in the symbol table is described below:

Average Type

RMS, Time

RMS, Time

Average Mode

any (single sweep)

Effects of averaging

After each scan, the Syms/Err table shows a running (linear) average over past scans for each parameter in the table. Peak or position parameters are not averaged. Parameters that appear in the table in dB are converted to linear units in order to average them. The measurement stops after the specified Avg

Number of scans.

Same as above, except that averages are reset after the specified Avg Number of scans, and the measurement continues.

RMS, Time

Max repeat

(continuous sweep) exponential

(continuous sweep) any

Same as the single sweep case until the specified Avg Number of scans is complete. After that, averaging continues using exponential weighting.

After each scan, compares each parameter in the table with the current scan's value and keeps the maximum. Symbol positions relate to the maximum peak value seen.

Average Mode

See

“Average Mode” on page 401

in the section "Common Measurement Functions 2" for a description of this function.

Average Setup

See

“Average Setup” on page 402

in the section "Common Measurement Functions 2" for a description of this function.

458

WLAN OFDM

Meas Setup

Demod Setup

This key brings up a menu of commonly used demodulation setup parameters.

Key Path

Mode

Meas Setup

VSA

Data Modulation Detect

This key allows you to select whether the data subcarrier modulation format is automatically detected or manually set to the particular modulation format selected by Modulation Format.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Range

[:SENSe]:W11A[:SUBCarrier]:DEMod:AUTO OFF|ON|0|1

[:SENSe]:W11A[:SUBCarrier]:DEMod:AUTO?

Set to Default by Preset to Standard.

W11A:DEM:AUTO ON

Meas Setup, Demod Setup

VSA

ON

Saved in instrument state.

Auto | Man

Modulation Format

This selects the Modulation format used when Data Modulation Detect is set to Manual.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Range

[:SENSe]:W11A[:SUBCarrier]:DEMod BPSK|QPSK|QAM16|QAM64

[:SENSe]:W11A[:SUBCarrier]:DEMod?

Set to Default by Preset to Standard.

W11A:DEM BPSK

Meas Setup, Demod Setup

VSA

BPSK

Saved in instrument state.

BPSK | QPSK | 16QAM | 64QAM

459

WLAN OFDM

Meas Setup

BPSK

Selects BPSK Data subcarrier Modulation format.

Key Path

Mode

Meas Setup, Demod Setup, Data Modulation

VSA

QAM

Selects QAM Data subcarrier Modulation format.

Key Path

Mode

Meas Setup, Demod Setup, Data Modulation

VSA

16QAM

Selects 16QAM Data subcarrier Modulation format.

Key Path

Mode

Meas Setup, Demod Setup, Data Modulation

VSA

64QAM

Selects 64QAM Data subcarrier Modulation format.

Key Path

Mode

Meas Setup, Demod Setup, Data Modulation

VSA

Guard Interval

The Guard Interval is the ratio of the Cyclic Prefix "CP" time to the inverse FFT time "T(IFFT)."

Remote Command

Coupling

Restriction and Notes

Example

Key Path

Mode

Preset

State Saved

Min

[:SENSe]:W11A:GINTerval <real>

[:SENSe]:W11A:GINTerval?

Set to Default by Preset to Standard.

Entries are rounded to the nearest 1/128.

W11A:GINT 0.25

Meas Setup, Demod Setup

VSA

0.25

Saved in instrument state.

0.0

460

WLAN OFDM

Meas Setup

Max 1.0

Preset to Standard

When one of the presets from the menu below this is selected, this immediately sets all measurement parameters to the appropriate values.

When "IEEE 802.11a/g OFDM" is chosen, this is the same as Meas Preset except that the display state is left alone.

In general, this sets all parameters to their Preset values. For the " IEEE 802.11a/g Turbo Mode" and "

IEEE 802.11p DSRC" presets choices, these parameters may be different: Span, Sync Training

Sequence, Subcarrier Spacing, and Search Length. For all presets, Standard is set. See the descriptions of these individual parameters for details.

About IEEE 802.11g modulation:

The IEEE 802.11g standard defines both single-carrier and OFDM modulation formats. The single-carrier modulation formats are compatible with 802.11b and can be demodulated using the VSA

WLAN DSSS measurement.

IEEE 802.11g defines two different OFDM modulation formats. One, which is mandatory in 802.11g, is an exact copy of the 802.11a format except that the carrier frequencies are in the 2.4 GHz band. We refer to this as "IEEE 802.11g OFDM". The other OFDM format is an optional format called DSSS-OFDM that combines an 802.11b-style single-carrier preamble with an 802.11a-style OFDM data payload. We refer to this as "IEEE 802.11g DSSS-OFDM". This measurement can demodulate and analyze either of these OFDM formats.

Remote Command

Example

Key Path

Mode

Preset

Range

[:SENSe]:W11A:STANdard:PRESet I11AGOFDM|HIPERLAN2|

I11GDSSSOFDM|I11AGTURBO|I11PDSRC|I11J10MHZ

W11A:STAN:PRES I11AGOFDM

Meas Setup

VSA

I11AGOFDM

IEEE 802.11a/g OFDM | HIPERLAN/2 | IEEE 802.11g DSSS-OFDM

| IEEE 802.11a/g Turbo Mode | IEEE 802.11p DSRC

| IEEE 802.11j 10 MHz

This table shows the mapping between the names of the selections on the softkey menu and their corresponding SCPI forms.

Preset soft key name

IEEE 802.11a/g OFDM

HIPERLAN/2

IEEE 802.1g DSSS-OFDM

Preset SCPI form

I11AGOFDM

HIPERLAN2

I11GDSSSOFDM

461

WLAN OFDM

Meas Setup

Preset soft key name

IEEE 802.11a/g Turbo Mode

IEEE 802.11p DSRC

IEEE 802.11j 10 MHz

Preset SCPI form

I11AGTURBO

I11PDSRC

I11J10MHZ

IEEE 802.11a/g OFDM

Sets the demodulation preset to IEEE std 802.11a–1999 specification.

Key Path

Mode

Meas Setup, Preset to Standard

VSA

HIPERLAN/2

Sets the demodulation preset to ETSI TS 101 475 V1.2.2 (2001–02) - Broadband Radio Access

Networks (BRAN); HIPERLAN Type 2; Physical (PHY) Layer specification.

Key Path

Mode

Meas Setup, Preset to Standard

VSA

IEEE 802.1g DSSS-OFDM

Sets the demodulation preset to IEEE std 802.11g – 2003 specification

Key Path

Mode

Meas Setup, Preset to Standard

VSA

IEEE 802.11a/g Turbo Mode

Sets the demodulation preset to a nonstandard double rate 802.11a.

Key Path

Mode

Meas Setup, Preset to Standard

VSA

IEEE 802.11p DSRC

Sets the demodulation preset to ASTM E2213 – 02 DSRC specification

Key Path

Mode

Meas Setup, Preset to Standard

VSA

462

WLAN OFDM

Meas Setup

IEEE 802.11j 10 MHz

Sets the demodulation preset to IEEE std 802.11j – 2004 specification

Key Path

Mode

Meas Setup, Preset to Standard

VSA

Meas Time

This key brings up a menu consisting of the time length and positioning demodulation setup parameters.

Key Path

Mode

Meas Setup

VSA

Search Length

Allows you to specify the length of time allowed to acquire the input signal (in seconds). This defines the length of time that the pulse search is performed.

The analyzer searches within the specified search length and demodulates the first complete pulse

(burst). The pulse can occur anywhere within the search length. Pulses beyond the first complete pulse are disregarded and are not demodulated. In order for the pulse search algorithm to detect a pulse edge, the pulse must be at least 15 dB above the noise floor. The resultant pulse is then demodulated.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11A:SYNC:SLENgth <time>

[:SENSe]:W11A:SYNC:SLENgth?

The minimum is constrained to meet the number of symbols of Result Length or Max Auto Result Length, whichever is active.

See the table below for the Preset to Standard values.

W11A:SYNC:SLEN 0.001

Meas Setup, Meas Time

VSA

1.0 ms

Saved in instrument state.

264 us

26.952 ms

The Search Length is set to the following by Preset to Standard:

Standard

IEEE 802.11a/g OFDM

Search Length

1.0 ms

463

WLAN OFDM

Meas Setup

Standard Search Length

HIPERLAN/2

IEEE 802.1g DSSS-OFDM

1.0 ms

1.0 ms

IEEE 802.11a/g Turbo Mode 0.5 ms

IEEE 802.11p DSRC 2.0 ms

IEEE 802.11j 10 MHz 2.0 ms

Result Length

This key allows you to either manually set the Result Length or to have it automatically detected.

The result length is the total number of symbols included in the acquired and demodulated data. The

Measurement Interval and Measurement Offset parameters are used to specify the measurement length within the Result Length.

When Result Length is Automatic, the result length is the number of symbols in the current burst, up to the limit of Max Auto Result Length.

Note that when Result Length is Manual, the demodulation is not limited to the number of symbols transmitted within the burst as is done when the Result Length is Auto. The analyzer will demodulate the total number of symbols specified by Result Length, which if longer than the burst, could lead to unexpected and incorrect data results.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11A:TIME:RESult:LENGth <integer>

[:SENSe]:W11A:TIME:RESult:LENGth?

[:SENSe]:W11A:TIME:RESult:AUTO OFF|ON|0|1

[:SENSe]:W11A:TIME:RESult:AUTO?

Search Length is increased to include the Result Length, if Result Length is in

Manual mode.

Set to Preset by Preset to Standard.

W11A:TIME:RES:LENG 60

W11A:TIME:RES:AUTO ON

Meas Setup, Meas Time

VSA

60 symbols

ON

Saved in instrument state.

1 symbol

1367 symbols

464

WLAN OFDM

Meas Setup

Max Auto Result Length

When Result Length is set to Auto this key causes the effective result length, the total number of symbols included in the demodulated data, to be automatically determined by comparing the input signal's length to the Max Auto Result Length value and using the smaller number.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11A:TIME:RESult:MAX <integer>

[:SENSe]:W11A:TIME:RESult:MAX ?

Search Length is increased to include the Max Auto Result Length, if Result

Length is in Auto mode.

Set to Default by Preset to Standard.

W11A:TIME:RESult:MAX 60

Meas Setup, Meas Time

VSA

60 symbols

Saved in instrument state.

1 symbol

1367 symbols

Meas Offset

Measurement Offset determines the start position, in symbols, of the Meas Interval with respect to the first symbol of the PLCP Preamble. The value must be less than the Result Length or Maximum Result

Length parameter (depending on which option is selected for Result Length). This parameter combined with Measurement Interval specifies the portion of the result length to analyze and display to the user.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11A:TIME:OFFSet <integer>

[:SENSe]:W11A:TIME:OFFSet?

The maximum is limited to Result Length or Max Auto Result Length, whichever is active, minus Meas Interval.

Set to Default by Preset to Standard.

W11A:TIME:OFFS 0

Meas Setup, Meas Time

VSA

0 symbols

Saved in instrument state.

0 symbols

See coupling

465

WLAN OFDM

Meas Setup

Meas Interval

Measurement Interval, in conjunction with Meas Offset, lets you specify an interval (segment) of the result length data to be demodulated and analyzed. Measurement Interval is entered as an integer number of Symbol-times.

The measurement interval value must be less than or equal to the Result Length or Maximum Result

Length parameter (depending on which option is selected for Result Length Select). This parameter combined with Measurement Offset specifies the portion of the result length to analyze and display to the user.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11A:TIME:INTerval <integer>

[:SENSe]:W11A:TIME:INTerval?

The maximum is limited to Result Length or Max Auto Result Length, whichever is active. If necessary, Meas Offset is reduced to keep Meas

Interval plus Meas Offset within Result Length or Max Result Length, whichever is active.

Set to Default by Preset to Standard.

W11A:TIME:INT 60

Meas Setup, Meas Time

VSA

60 symbols

Saved in instrument state.

1 symbol

See coupling

Subcarriers

Allows you to specify what Subcarrier data are analyzed and data results displayed.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Range

[:SENSe]:W11A:SUBCarrier:SELect ALL|PILot|SINGle

[:SENSe]:W11A:SUBCarrier:SELect?

Set to Default by Preset to Standard.

W11A:SUBC:SEL ALL

Meas Setup, More

VSA

ALL

Saved in instrument state.

All | Pilots | Single

466

WLAN OFDM

Meas Setup

All

Displays data results for all subcarriers. (–26 Through 26)

Key Path

Mode

Meas Setup, More, Subcarriers

VSA

Pilots

Displays data results for all Pilot subcarriers. Pilots subcarriers include subcarrier –21, –7, 7, and 21.

Key Path

Mode

Meas Setup, More, Subcarriers

VSA

Single

Displays data results for the selected Subcarrier.

Key Path

Mode

Meas Setup, More, Subcarriers

VSA

Subcarrier

Selects the single subcarrier that is used when Single is chosen.

Remote Command

Coupling

Restriction and Notes

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11A:SUBCarrier:NUMBer <integer>

[:SENSe]:W11A:SUBCarrier:NUMBer?

Set to Default by Preset to Standard.

Subcarrier 0 is unused and so may not be selected.

W11A:SUBC:NUMB 1

Meas Setup, More, Subcarriers, Single

VSA

1

Saved in instrument state.

–26

26

Advanced

This key displays a menu that allows you to select lesser used demodulation parameters for the current measurement. These settings are for advanced users and do not normally require adjustment for most

467

WLAN OFDM

Meas Setup

common measurements.

Key Path

Mode

Meas Setup, More

VSA

Subcarrier Spacing

Set the Subcarrier spacing.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11A:SUBCarrier:SPACing <freq>

[:SENSe]:W11A:SUBCarrier:SPACing?

Span is forced to be at least 5.3 times the Subcarrier Spacing. So the maximum Span sets the maximum Subcarrier Spacing. Likewise, the Span will be forced to be within 530 times the Subcarrier Spacing and so the minimum Span sets the minimum Subcarrier Spacing.

See the table below for the Preset to Standard values.

W11A:SUBC:SPAC 312.5E3

Meas Setup, More, Advanced

VSA

312.5 kHz

Saved in instrument state.

See Coupling

See Coupling

The Subcarrier Spacing is set to the following by Preset to Standard:

Standard Subcarrier Spacing

IEEE 802.11a/g OFDM

HIPERLAN/2

312.5 kHz

312.5 kHz

IEEE 802.1g DSSS-OFDM 312.5 kHz

IEEE 802.11a/g Turbo Mode 625 kHz

IEEE 802.11p DSRC

IEEE 802.11j 10 MHz

156.25 kHz

156.25 kHz

Symbol Timing Adjust

Shifts the start of the TFFT period earlier in the symbol. You specify the amount of TFFT shift as a percentage of the TFFT length.

Moving the TFFT period away from the ending transition time (TTR) into the guard interval (TGI) may

468

WLAN OFDM

Meas Setup

help avoid inter-symbol interference. Make sure that the TFFT period is not shifted so much that it includes corrupt data from the transition time at the beginning of the symbol.

Symbol Timing Adjust allows you to adjust the "useful symbol time period" (TFFT) within the "OFDM extended symbol time period" (TS). Symbol Timing Adjust shifts the start of the TFFT period earlier in the TS time period. You specify the amount of TFFT shift as a percentage of the TFFT length.

The extended OFDM symbol time period (TS) consists of a guard interval (TGI) plus a "useful symbol time period" (TFFT). Within the TS, the analyzer performs the demodulation and data analysis on only one TFFT time period. The Symbol Time Adjust parameter allows you to move the TFFT within the TS.

This value is expressed as a percentage of the TFFT length.

Typically, the TGI time period is ingnored and only the TFFT time period is used. The Symbol Time

Adjust parameter causes the demodulation start time to begin earlier within each extended symbol time.

The minimum Symbol Time Adjust value is equal to "guard interval*100", the maximum value is "0"

(full guard interval).

T

GI

T =T +T

S GI FFT

T

FFT

T moved by

T

STA

T

FFT

T

TR

T

STA

(%)

T

GI

= Guard Interval

T = Symbol Transition Time

TR

T

STA

= Symbol Timing Adjust (%)

OFDM Frame with cyclic extension for a single symbol time

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

Min

Max

[:SENSe]:W11A:TADJust <real>

[:SENSe]:W11A:TADJust?

Set to Default value for all Standards by Preset to Standard.

W11A:TADJ –3.125

Meas Setup, More, Advanced

VSA

–3.125 %

Saved in instrument state.

–25 %

0.0 %

469

WLAN OFDM

Meas Setup

Pilot Tracking

This key displays a menu of several Pilot Tracking setup parameters.

Key Path

Mode

Meas Setup, More, Advanced

VSA

Track Amplitude

When Track Amplitude is selected, the analyzer applies pilot subcarrier amplitude error correction to the pilot and data subcarriers. This is in addition to Track Phase and Track Timing error correction if selected.

This parameter specifies whether the analyzer tracks amplitude changes in the pilot subcarriers. When true, the Common Pilot Error trace will show the amplitude tracking in the magnitude portion of the trace

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

[:SENSe]:W11A:TRACk:AMPLitude OFF|ON|0|1

[:SENSe]:W11A:TRACk:AMPLitude?

Set to Default by Preset to Standard.

W11A:TRAC:AMPL OFF

Meas Setup, More, Advanced, Pilot Tracking

VSA

OFF

Saved in instrument state.

Track Phase

When Track Phase is selected, the analyzer applies pilot subcarrier phase error correction to the pilot and data subcarriers. This is in addition to Track Amplitude and Track Timing error correction if selected.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

[:SENSe]:W11A:TRACk:PHASe OFF|ON|0|1

[:SENSe]:W11A:TRACk:PHASe?

Set to Default by Preset to Standard.

W11A:TRAC:PHASe ON

Meas Setup, More, Advanced, Pilot Tracking

VSA

ON

Saved in instrument state.

Track Timing

When Track Timing is selected the analyzer applies pilot subcarrier timing error correction (frequency

470

WLAN OFDM

Meas Setup

offset correction) to the pilot and data subcarriers. This is in addition to Track Amplitude and Track

Phase error correction if selected.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

[:SENSe]:W11A:TRACk:TIMing OFF|ON|0|1

[:SENSe]:W11A:TRACk:TIMing?

Set to Default by Preset to Standard.

W11A:TRAC:TIM OFF

Meas Setup, More, Advanced, Pilot Tracking

VSA

OFF

Saved in instrument state.

EQ Training

When demodulating an OFDM signal, the analyzer uses an equalizer to correct for linear impairments in the signal path, such as multi-path. The 89600 analyzer supports two different ways to initialize, or

"train," the equalizer.

The valid values are Channel Estimation Seq Only and Channel Estimation Seq & Data. The value

Channel Estimation Seq Only, which is the default, specifies that the equalizer is trained using only the channel estimation sequence (also called the "long sync") portion of the burst preamble.

The value and Channel Estimation Seq & Data specifies that the equalizer should be trained using both the channel estimation sequence and the entire data portion of the burst. This usually gives a more accurate estimate of the equalizer response. It also typically lowers the EVM by one or two dB.

Remote Command

Coupling

Example

Key Path

Mode

Preset

Range

[:SENSe]:W11A:EQUalizer:TRAining CHANnel|CDATa

[:SENSe]:W11A:EQUalizer:TRAining?

Set to Default by Preset to Standard.

W11A:EQU:TRA CHAN

Meas Setup, More, Advanced

VSA

Channel Est Seq Only

Channel Est Seq Only | Channel Est Seq & Data

Channel Est Seq Only

The equalizer is trained by looking at the Channel Estimation Sequence in the preamble of the OFDM burst. After this initialization, the equalizer coefficients are held constant while demodulating the rest of the burst.

471

WLAN OFDM

Meas Setup

Downlink Subframe

P H

B1 B2

Synchronization

Sequence

Channel

Estimation

Sequence

Symbol 1 Symbol 2

Long Preamble

Key Path

Mode

Meas Setup, More, Advanced, EQ Training

VSA

Channel Est Seq & Data

The equalizer is trained by analyzing the entire OFDM burst, including the Channel Estimation

Sequence (contained in the preamble) and the Data symbols. This type of training generally gives a more accurate estimate of the true response of the transmission channel.

Key Path

Mode

Meas Setup, More, Advanced, EQ Training

VSA

Sync Training Sequence

The Sync parameters control the type of synchronization used to synchronize to the start of an OFDM burst. The two available choices are "Short Training Sequence" (also called Short Sync or Short Training

Symbols) and "Chan Estimation Seq" (also called Long Sync or Long Training Symbols).

Remote Command

Coupling

Example

Key Path

Mode

Preset

Range

[:SENSe]:W11A:SYNC:SEQuence LONG|SHORt

[:SENSe]:W11A:SYNC:SEQuence?

See the table below for the Preset to Standard values.

W11A:SYNC:SEQ SHOR

Meas Setup, More, Advanced

VSA

SHOR

Long | Short

472

WLAN OFDM

Meas Setup

The Sync Training Seq is set to the following by Preset to Standard:

Standard Sync Training Seq

IEEE 802.11a/g OFDM

HIPERLAN/2

Short

Long (Channel Estimation)

IEEE 802.1g DSSS-OFDM Long (Channel Estimation)

IEEE 802.11a/g Turbo Mode Long (Channel Estimation)

IEEE 802.11p DSRC

IEEE 802.11j 10 MHz

Short

Short

IQ Normalize

When set to On, the IQ trace data results (including IQ Meas, IQ Ref, Error Vector Time and corresponding Spectrum results) are normalized.

When set to Off, the trace data results are not normalized.

When normalization is On, the analyzer normalizes or scales the demodulated trace data results, in magnitude, to a nominal value of 1, relative to the outermost states of the ideal (IQ Reference) constellation diagram.

Note that regardless of this setting, the Error Summary trace data results are always normalized.

Remote Command

Coupling

Example

Key Path

Mode

Preset

State Saved

:CALCulate:W11A:NORMalize OFF|ON|0|1

:CALCulate:W11A:NORMalize?

Set to Default by Preset to Standard.

CALC:W11A:NORM ON

Meas Setup, More, Advanced

VSA

ON

Saved in instrument state.

Standard

This single parameter takes care of setting a number of internal, standard specific details of the demodulation. The valid values are the same as that of Preset to Standard and are set to the same by

Preset to Standard.

Remote Command

Coupling

[:SENSe]:W11A:STANdard I11AGOFDM | HIPERLAN2 |

I11GDSSSOFDM | I11AGTURBO | I11PDSRC | I11J10MHZ

[:SENSe]:W11A:STANdard?

Set to matching Standard by Preset to Standard.

473

WLAN OFDM

Meas Setup

Example

Key Path

Mode

Preset

Range

W11A:STAN I11J10MHZ

Meas Setup, More, Advanced

VSA

IEEE 802.11a/g OFDM

IEEE 802.11a/g OFDM | HIPERLAN/2 | IEEE 802.1g DSSS-OFDM

| IEEE 802.11a/g Turbo Mode | IEEE 802.11p DSRC

| IEEE 802.11j 10 MHz

This table shows the mapping between the names of the selections on the softkey menu and their corresponding SCPI forms.

Preset soft key name

IEEE 802.11a/g OFDM

HIPERLAN/2

IEEE 802.1g DSSS-OFDM

IEEE 802.11a/g Turbo Mode

IEEE 802.11p DSRC

IEEE 802.11j 10 MHz

Preset SCPI form

I11AGOFDM

HIPERLAN2

I11GDSSSOFDM

I11AGTURBO

I11PDSRC

I11J10MHZ

IEEE 802.11a/g OFDM

Per the IEEE std 802.11a–1999 specification.

Key Path

Mode

Meas Setup, Preset to Standard

VSA

HIPERLAN/2

Per the ETSI TS 101 475 V1.2.2 (2001–02) - Broadband Radio Access Networks (BRAN); HIPERLAN

Type 2; Physical (PHY) Layer specification.

Key Path

Mode

Meas Setup, Preset to Standard

VSA

IEEE 802.1g DSSS-OFDM

Per the IEEE std 802.11g – 2003 specification

Key Path

Mode

Meas Setup, Preset to Standard

VSA

474

WLAN OFDM

Meas Setup

IEEE 802.11a/g Turbo Mode

This is a nonstandard double rate 802.11a.

Key Path

Mode

Meas Setup, Preset to Standard

VSA

IEEE 802.11p DSRC

Per the ASTM E2213 – 02 DSRC specification

Key Path

Mode

Meas Setup, Preset to Standard

VSA

IEEE 802.11j 10 MHz

Per the IEEE std 802.11j – 2004 specification

Key Path

Mode

Meas Setup, Preset to Standard

VSA

Meas Preset

This immediately sets all measurement parameters to their 802.11a/g OFDM Preset values. This presets the display state in the same way as Preset View: Basic.

For more information, see the section under the Preset key in the Utility section.

Key Path

Mode

Meas Setup, More

VSA

475

WLAN OFDM

Mode

Mode

See

“Mode” on page 213 in the section "Common Measurement Functions" for a description of this

function.

476

WLAN OFDM

Mode Setup

Mode Setup

See “Mode Setup” on page 229 in the section "Common Measurement Functions" for a description of

this function.

477

WLAN OFDM

Peak Search

Peak Search

See

“Peak Search” on page 406

in the section "Common Measurement Functions 2" for a description of this function.

478

WLAN OFDM

Recall

Recall

See “Recall” on page 139 in the section "System Functions" for a description of this function.

State

See “State” on page 139

in the section "System Functions" for a description of this function.

Data

See “Data (Import)” on page 146 in the section "System Functions" for a description of this function.

Trace

See “Import Trace Data” on page 411

in the section "Common Measurement Functions 2" for a description of this function..

Open

See “Open…” on page 150 in the section "Common Measurement Functions 2" for a description of this

function.

479

WLAN OFDM

Restart

Restart

See

“Restart” on page 258 in the section "Common Measurement Functions" for a description of this

function.

480

WLAN OFDM

Save

Save

See “Save” on page 155 in the section "System Functions" for a description of this function.

State

See “State” on page 155

in the section "System Functions" for a description of this function.

Data

See “Data (Export)” on page 162 in the section "System Functions" for a description of this function.

Trace

See “Export Trace Data” on page 413 in the section "Common Measurement Functions 2" for a

description of this function.

Save As…

See “Save As . . .” on page 165 in the section "System Functions" for a description of this function.

Screen Image

See “Screen Image” on page 166 in the section "System Functions" for a description of this function.

481

WLAN OFDM

Single

Single

See

“Single (Single Measurement/Sweep)” on page 263 in the section "Common Measurement

Functions" for a description of this function.

482

WLAN OFDM

Source

Source

See “Source” on page 263

in the section "Common Measurement Functions" for a description of this function.

483

WLAN OFDM

SPAN X Scale

SPAN X Scale

See

“SPAN X Scale” on page 416 in the section "Common Measurement Functions 2" for a description

of this function.

Span

This controls the frequency span of the measurement. See “FREQ Channel” on page 371 in the section

“Common Measurement Functions 2” for details on how this interacts with Start, Stop, and Center frequencies.

Coupling

Key Path

Mode

Preset

Limited to be within a factor of 5.3 to 530 of the Subcarrier Spacing.

When Subcarrier Spacing is changed, Span may be changed to meet this requirement.

See the table below for the Preset to Standard values.

SPAN X Scale

VSA

31.25 MHz (limited by hardware to 10 MHz or 25 MHz with Option B25)

The Span is set to the following by Preset to Standard:

Standard Span

IEEE 802.11a/g OFDM

HIPERLAN/2

31.25 MHz

31.25 MHz

IEEE 802.1g DSSS-OFDM 31.25 MHz

IEEE 802.11a/g Turbo Mode 62.5 MHz

IEEE 802.11p DSRC

IEEE 802.11j 10 MHz

15.625 MHz

15.625 MHz

Full Span

See Span X Scale,

“Full Span” on page 416 in the section "Common Measurement Functions 2" for a

description of this function.

Select Trace

See Trace/Detector,

“Select Trace” on page 428

in the section "Common Measurement Functions 2" for a description of this function.

484

WLAN OFDM

SPAN X Scale

X Scale

See Span X Scale, “X Scale” on page 420

in the section "Common Measurement Functions 2" for a description of this function.

Freq Annotation

See Span X Scale, “Freq Annotation” on page 423

in the section "Common Measurement Functions 2" for a description of this function.

All Frequency Points

See Span X Scale,

“All Frequency Points” on page 422

in the section "Common Measurement Functions

2" for a description of this function.

Copy X Scale

See Span X Scale, “Copy X Scale” on page 423 in the section "Common Measurement Functions 2" for

a description of this function.

485

WLAN OFDM

Sweep/Control

Sweep/Control

See

“Sweep/Control” on page 424

in the section "Common Measurement Functions 2" for a description of this function.

486

WLAN OFDM

Trace/Detector

Trace/Detector

This section details the trace results accessible via the data key and also via SCPI, many of which are available in other measurements as well.

See “Trace/Detector” on page 428 in the section "Common Measurement Functions" for a description of

this function.

Select Trace

See “Select Trace” on page 428 in the section "Common Measurement Functions 2" for a description of

this function.

Data

This key provides a menu of trace data choices for the selected trace. For the SCPI command, and other details, see Trace/Detector,

“Data” on page 430

.

The following are the SCPI string forms for the Data types that are specific to this measurement, in alphabetical order:

Trace data soft key name

CCDF

CDF

Ch Frequency Response

Common Pilot Error

Eq Impulse Response

Error Vector Spectrum

Error Vector Time

IQ Meas

IQ Ref

Inst Ch Frequency Response

Inst Spectrum

PDF

Preamble Error

Preamble Freq Error

RMS Error Vector Spectrum

RMS Error Vector Time

SCPI string form

"CCDF1"

"CDF1"

"Ch Frequency Response1"

"Common Pilot Error1"

"Eq Impulse Response1"

"Error Vector Specrum1"

"Error Vector Time1"

"IQ Meas1"

"IQ Ref1"

"Inst Ch Frequency Response1"

"Inst Spectrum1"

"PDF1"

"Preamble Error1"

Preamble Freq Error1

RMS Error Vector Spectrum1

RMS Error Vector Time1

487

WLAN OFDM

Trace/Detector

Trace data soft key name

Raw Main Time

Search Time

Spectrum

Symbols/Errors

Time

SCPI string form

"Raw Main Time1"

"Search Time1"

"Spectrum1"

"Syms/Errs1"

"Time1"

Key Path

Mode

Trace/Detector

VSA

Pre Demod

This key accesses the Trace Data choices which show pre-demodulation results.

Key Path

Mode

Trace/Detector, Data

VSA

Spectrum

Averaged FFT of the Time waveform. See Trace/Detector, Data, Pre-demod, “Spectrum” on page 431 in

the section "Common Measurement Functions 2" for more information.

Inst Spectrum

The FFT of the time waveform for the current measurement. “Inst” or Instantaneous refers to this result not being averaged like the Spectrum Trace Data result. See Trace/Detector, Data, Pre-demod,

“Inst

Spectrum” on page 433

in the section "Common Measurement Functions 2" for more information.

Search Time

The Search Length long time record acquired for the current measurement.

Key Path

Mode

Trace/Detector, Data, Pre Demod

VSA

Time

The time record before digital demodulation and after the pulse/burst search.

The starting point is the beginning of the burst, offset by the Meas Offset and its length is the Meas

Interval.

The exception to this is that 10% more additional data on both sides is shown when the Measurement

Offset is set to zero and the Measurement Interval is set the same as the Result Length in Manual Result

Length mode. In automatic Result Length mode, this will occur when the Measurement Offset to zero and the Measurement Interval is set to the Max Auto Result Length, but will not go more than 10%

488

WLAN OFDM

Trace/Detector

beyond the end of the burst.

Key Path

Mode

Raw Main Time

The raw time record acquired for the current measurement. This data is unprocessed and includes additional points acquired for settling of the filters involved in subsequent processing, such as the

demodulation filtering. See Trace/Detector, Data, Pre-demod, “Raw Main Time” on page 436

in the section "Common Measurement Functions 2" for more information.

Statistical

This key accesses the Trace Data choices which show statistical results.

Key Path

Mode

Trace/Detector, Data, Pre Demod

VSA

Trace/Detector, Data

VSA

CCDF

The Complementary, Cumulative Density function (CCDF) for the selected input channel.

The analyzer plots CCDF using units of percent (%) for the y-axis and power (dB) for the x-axis. Power

on the x-axis is relative to the signal average power. See Trace/Detector, Data, Statistical, “CCDF

(Complementary, Cumulative Density Function)” on page 441 in the section "Common Measurement

Functions 2" for more information.

CDF

The Cumulative Density Function (CDF) for the selected input channel. CDF is computed by integrating the CCDF (Probability Density Function). See Trace/Detector, Data, Statistical,

“CDF (Cumulative

Density Function)” on page 442

in the section "Common Measurement Functions 2" for more information.

PDF

The Probability Density Function (PDF) for the selected input channel. PDF indicates the probability that a given level has occurred. See Trace/Detector, Data, Statistical,

“PDF (Probability Density

Function)” on page 442 in the section "Common Measurement Functions 2" for more information.

Demod

This key accesses the Trace Data choices which show general demodulation results.

Key Path

Mode

Trace/Detector, Data

VSA

IQ Meas

IQ Meas is the measured IQ symbol values of the subcarriers. There is one complex value for each

489

WLAN OFDM

Trace/Detector

subcarrier for each symbol in the burst.

Normally this trace data is displayed as a constellation. The constellation display shows both data and pilot subcarriers, the pilots and data values are shown with different colors.

With most other display formats, the data is plotted vs. subcarrier, with the points for each symbol all plotted separately. This is the same kind of display as Error Vector Spectrum.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

IQ Ref

IQ Ref is the reference (ideal) IQ values of the subcarriers. There is one complex value for each subcarrier for each symbol in the burst.

Normally this trace data is displayed as a constellation. The constellation shows both data and pilot subcarrier symbols, the pilots and data values are shown with different colors.

With most other display formats, the data is plotted vs. subcarrier, with the points for each symbol all plotted separately. This is the same kind of display as Error Vector Spectrum.

Key Path

Mode

Trace/Detector, Data, Demod

VSA

Demod Error

This key accesses the Trace Data choices which show demodulation error related results.

Key Path

Mode

Trace/Detector, Data

VSA

Error Vector Time

This trace, like Error Vector Spectrum shows each of the individual signal error vectors for each subcarrier and symbol vs. Time (symbol) and frequency (subcarrier). Each error vector is the vector difference, for that subcarrier at that symbol-time, between the corresponding IQ Meas value and the IQ

Ref value.

On this trace, the individual error vectors are plotted vs Time (symbol). So at each valid symbol, there is a point plotted for each valid subcarrier (52 total, since subcarrier 0 is not used.) In addition, a white trace is drawn, where each point is the RMS average over the valid subcarriers, which is the same result as is plotted separately as RMS Error Vector Time.

Key Path

Mode

Trace/Detector, Data, Demod Error

VSA

RMS Error Vector Time

RMS Error Vector Time is the difference between IQ Meas and IQ Ref is the error vector (which would

490

WLAN OFDM

Trace/Detector

have a complex value) at each subcarrier at each symbol-time. This trace is the RMS average of the error vector for each valid subcarrier at the plotted symbol, the same data shown as a white trace shown in Error Vector Time.

Key Path

Mode

Trace/Detector, Data, Demod Error

VSA

Error Vector Spectrum

This trace, like Error Vector Time shows each of the individual signal error vectors for each subcarrier and symbol vs. Time (symbol) and frequency (subcarrier). Each error vector is the vector difference, for that subcarrier at that symbol-time, between the corresponding IQ Meas value and the IQ Ref value.

On this trace, the individual error vectors are plotted vs frequency (subcarrier). So at each valid subcarrier, there is a point plotted for each valid symbol. Note that subcarrier 0 is not plotted since it is not used. In addition, a white trace is drawn, where each point is the RMS average over the valid symbols, which is the same result as is plotted separately as RMS Error Vector Spectrum.

Key Path

Mode

Trace/Detector, Data, Demod Error

VSA

RMS Error Vector Spectrum

This trace is the RMS average of the error vector for each valid symbol at the plotted subcarrier, the same data shown as a white trace shown in Error Vector Time. Note that subcarrier 0 is not plotted since it is not used.

Key Path

Mode

Trace/Detector, Data, Demod Error

VSA

Common Pilot Error

This trace shows the difference between the measured and ideal pilot subcarrier symbols.

At each symbol in the burst, the measured symbol values of the four pilot subcarriers are compared with the ideal values. The differences are averaged together, producing a single complex value for each symbol. Normally the phase of this trace is displayed, showing how the pilot phase changes over the burst. You can also look at the magnitude of this trace to see signal magnitude changes over the burst.

To display in frequency units, select the Group Delay Trace Format. See Analyzer Setup Functions,

Trace/Detector, Format.

Key Path

Mode

Trace/Detector, Data, Demod Error

VSA

Preamble Error

Preamble Error shows the correlation between segments of the measured preamble signal with the ideal preamble signal. You can select the desired preamble sequence to be compared to with the EQ Training

491

WLAN OFDM

Trace/Detector

and Sync Training Sequence parameters.

Key Path

Mode

Trace/Detector, Data, Demod Error

VSA

Preamble Freq Error

Preamble Frequency error is the difference between the measured center frequency of the transmitted signal and the measurement center frequency. The Preamble Frequency Error trace shows the total frequency error during the preamble portion of the OFDM burst. This includes the constant frequency error as displayed in the Symbols/Errors trace in addition to any time-varying frequency error.

When Sync Training Sequence is set to Short (the default for 802.11a), Preamble Frequency Error covers both the short training sequence and the channel estimation sequence. When the Sync type parameter is set to Chan Est, Preamble Frequency Error covers only the channel estimation sequence.

Key Path

Mode

Trace/Detector, Data, Demod Error

VSA

Tables

This key accesses the Trace Data choices which are in tabular form, including demodulated symbols tables.

Key Path

Mode

Trace/Detector, Data

VSA

Symbols/Errors

This display contains an error summary in the upper part, and a listing of demodulated symbols in the lower part. The error summary consists of:

Result name

EVM (dB)

EVM (% rms)

Pilot EVM

Common Pilot Error

Frequency Error

IQ Offset

Quadrature Error

Gain Imbalance

Sync Correlation

Displayed

Unit

Hz dB deg dB dB

%rms dB

%rms

Remote Name

EVMdB

EVM

PilotEVM

CPE

FreqErr

IQOffset

IQQuadErr

IQGainImb

SyncCorr

Remote

Unit

dB

%rms dB

%rms

Hz dB deg dB

(none)

492

WLAN OFDM

Trace/Detector

Result name

Symbol Clock Err

Modulation Format

Octets

Symbols

Code Rate

Bit Rate

Displayed

Unit

ppm

Remote Name

SymClkErr

ModFmt

Octets

Syms

CodeRate

BitRate

Remote

Unit

(none)

(none)

(none) symbols

(none) bps bps

The error summary values may be obtained using the CALC:W11A:DATA:TABL commands.

The demodulated symbols are available as the data values for this trace and so may be accessed using the

:CALCulate:W11A:DATA command. See Common Functions, Data Queries, CALCulate:DATA in the

VSA Mode PD for more details.

Key Path

Mode

Trace/Detector, Data, Tables

VSA

Response

This key accesses the Trace Data choices which show equalizer response results.

Key Path

Mode

Trace/Detector, Data

VSA

Ch Frequency Response

Ch Frequency Response shows the equalizer channel frequency response, which is the reciprocal of the equalizer frequency response. This is the measured equalizer frequency response for the burst, based on analyzing the preamble of the burst. It contains one complex value for each subcarrier, plus an interpolated value at the middle unused subcarrier (a total of 53 values).

This trace is averaged when Averaging is turned on.

Key Path

Mode

Trace/Detector, Data, Response

VSA

Inst Ch Frequency Resp

This is the instantaneous (non-averaged) Ch Frequency Response. If averaging is on, the Inst Ch

Frequency Response does not show the effects of averaging.

Key Path

Mode

Trace/Detector, Data, Response

VSA

493

WLAN OFDM

Trace/Detector

EQ Impulse Response

This shows the impulse response of the equalization filter. The equalizer impulse response is computed by taking the reciprocal of the channel equalizer frequency response, performing data filtering and computations that produce a result length of 4x the FFT length, and then converting to the time domain.

The Eq Impulse Response is the computed channel impulse response used to compensate for signal channel response degradation.

Key Path

Mode

Trace/Detector, Data, Response

VSA

ACP

This brings up a selection of the ACP result traces. See Trace/Detector, Data, “ACP Setup” on page 453

in the section "Common Measurement Functions 2" for more information.

OBW

This brings up a selection of the OBW result traces. See Trace/Detector, Data,

“OBW Setup (Occupied

Bandwidth)” on page 463 in the section "Common Measurement Functions 2" for more information.

Register

This brings up a selection of the Data Registers. See Trace/Detector, Data,

“Register” on page 468

in the section "Common Measurement Functions 2" for more information.

No Data

A blank display is shown.

Format

See

“Format” on page 443 in the section "Common Measurement Functions 2" for a description of this

function.

Digital Demod Trace Setup

See

“Digital Demod Trace Setup” on page 444

in the section "Common Measurement Functions 2" for a description of this function.

Copy to Data Register

See

“Copy to Data Register” on page 450

in the section "Common Measurement Functions 2" for a description of this function.

Phase/Delay Properties

See

“Phase/Delay Properties” on page 451 in the section "Common Measurement Functions 2" for a

description of this function.

494

WLAN OFDM

Trace/Detector

ACP Setup

See “ACP Setup” on page 453

in the section "Common Measurement Functions 2" for a description of this function.

OBW Setup

See “OBW Setup (Occupied Bandwidth)” on page 463

in the section "Common Measurement Functions

2" for a description of this function.

Trace Indicator Info

See “Trace Indicator Info” on page 469 in the section "Common Measurement Functions 2" for a

description of this function.

495

WLAN OFDM

Trigger

Trigger

See

“Trigger” on page 470

in the section "Common Measurement Functions 2" for a description of this function.

496

WLAN OFDM

View/Display

View/Display

See “View/Display” on page 479

in the section "Common Measurement Functions 2" for a description of this function.

Display

See “Display” on page 349 in the section "Common Measurement Functions" for a description of this

function.

Layout

See “Layout” on page 479

in the section "Common Measurement Functions 2" for a description of this function.

Preset View

Remote Command

Example

Key Path

Mode

:DISPlay:W11A:VIEW:PRESet BASic|EVM|POWer|DIAGnostic

DISP:W11A:VIEW:PRES BAS

(SCPI only)

VSA

Preset View: Basic

This preset view consists of the following traces in a Grid 2x2 layout:

1. IQ Meas Time in I-Q format

2. Spectrum in Log Mag (dB) format

3. Search Time in Log Mag (dB) format

4. Symbols/Errors

The Preset View: Basic softkey does an immediate action of changing the layout and view to this state.

This Preset View is an action, not a state.

This layout is the layout set by Meas Preset and is good for insuring that the signal is being demodulated correctly, as well as giving visibility to many basic demodulation setup problems.

Remote Command

Example

Key Path

Mode

:DISPlay:W11A:VIEW:PRESet BASic

DISP:W11A:VIEW:PRES BAS

View/Display

VSA

497

WLAN OFDM

View/Display

Preset View: EVM

This preset view consists of the following traces in a Stacked layout:

1. Error Vector Spectrum in Linear Mag format

2. Symbols/Errors

This layout is for more detailed EVM analysis.

The Preset View: EVM softkey does an immediate action of changing the layout and view to this state.

This Preset View is an action, not a state.

Remote Command

Example

Key Path

Mode

:DISPlay:W11A:VIEW:PRESet EVM

DISP:W11A:VIEW:PRES EVM

View/Display

VSA

498

WLAN OFDM

View/Display

Preset View: Power

This preset view consists of the following traces in a Grid 2x2 layout:

1. Spectrum in Log Mag (dB) format

2. Time in Log Mag (dB) format

3. Search Time in Log Mag (dB) format

4. CCDF in Log Mag (Linear Unit) format

This layout is oriented towards examining the signal in the Time and Power domain.

The Preset View: Basic softkey does an immediate action of changing the layout and view to this state.

This Preset View is an action, not a state.

Remote Command

Example

Key Path

:DISPlay:W11A:VIEW:PRESet POWer

DISP:W11A:VIEW:PRES POW

View/Display

499

WLAN OFDM

View/Display

Mode VSA

Preset View: Diagnostic

This preset view consists of the following traces in a Grid 2x2 layout:

1. Error Vector Spectrum in Linear Mag format

2. Ch Frequency Response in Log Mag (dB) format

3. Error Vector Time in Linear Mag format

4. Common Pilot Error in Wrap Phase format

This layout is useful for diagnosing modulation impairments and errors.

The Preset View: Basic softkey does an immediate action of changing the layout and view to this state.

This Preset View is an action, not a state.

Remote Command

Example

:DISPlay:W11A:VIEW:PRESet DIAGnostic

DISP:W11A:VIEW:PRES DIAGnostic

500

Key Path

Mode

View/Display

VSA

WLAN OFDM

View/Display

501

WLAN OFDM

View/Display

502

12

Common Measurement Functions

The key and command descriptions in this section describe functions that operate the same in multiple measurements and/or modes. This section is a library of functions that is referenced by many measurements and modes

To find the exact description and parameters for functions in a specific measurement, always look in the measurement section of this documentation. Pressing the front-panel key or softkey and then pressing the green Help key also provides the correct information.

NOTE

If you want to print the documentation, be sure to select this section and the measurement of interest to ensure having all the information you need. See

“Printing Acrobat Files” on page 81

for further instructions about printing.

503

Common Measurement Functions

504 Chapter 12

Common Measurement Functions 1

Common Measurement Functions 1

The key and command descriptions in this section describe functions that operate identically in multiple measurements and/or modes. This section is a library of functions that is referenced by many measurements and modes.

measurement section of this documentation. Pressing the front-panel key or softkey and then pressing the green Help key also provides the correct information.

NOTE

If you want to print the documentation, be sure to select this section and the measurement of interest to ensure having all the information you need. See

“Printing Acrobat Files” on page 145 for further instructions about printing.

AMPTD Y Scale

The Amplitude front-panel key activates the Amplitude menu and selects Reference Level or Reference

Value (depending on the measurement) as the active function.

Some features in the Amplitude menu apply to multiple measurements; others apply only to specific measurements. Keys that only apply to some measurements are blanked or grayed out in measurements in which they are not supported.

Key Path

Initial S/W Revision

Front-panel key

Prior to A.02.00

Reference Level

The Reference Level specifies the amplitude represented by the topmost graticule line.

Changing the reference level does not restart a measurement, because it is a display function only; instead it vertically ‘pans’ all displayed traces and markers to the new value. If a change to the reference level changes the attenuation value (eg through an auto coupling), then the measurement will be restarted.

Key Path

Remote Command

Example

AMPTD Y Scale

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel <real>

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel?

DISP:WIND:TRAC:Y:RLEV 20 dBm

Sets the reference level to 20 dBm, which displays in the current Y axis unit.

For example, if the Y axis unit is dB

μV, then 126.99 dBμV will be displayed.

1

Common Measurement Functions 1

Couplings

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

If you reduce the attenuation, the analyzer may have to lower the reference level to keep it below its allowed maximum. This allowed maximum level is specified in the “Max” row, below, along with other variables which affect it.

When you increase attenuation, the reference level does not change.

0 dBm

Saved in instrument state

RefLevelMin = –170 dBm + RefLevelOffset - ExtGain.

The maximum Ref Level is typically:

+30 dBm + RL Offset – External Gain (for MXA and PXA)

+23 dBm + RL Offset – External Gain (for EXA and CXA)

This maximum value is determined by the maximum power that can be safely applied to the input circuitry. The actual maximum value at any given time may be even less than this, depending on other values including Mech Atten,

Int Preamp Gain, Swept IF Gain, FFT IF Gain, Max Mixer Level, and the total attenuation currently available.

Depends on the current selected Y axis unit

Prior to A.02.00

Amplitude Representations

The following is an illustration of the reference level and Y Axis scales under various conditions:

Attenuation

This menu controls the attenuator functions and interactions between the attenuation system components.

There are two attenuator configurations in the X-Series. One is a dual attenuator configuration consisting of a mechanical attenuator and an optional electronic attenuator. The other configuration uses a single attenuator with combined mechanical and electronic sections thatcontrols all the attenuation functions. Different models in the X-Series come with different configurations.

2

Common Measurement Functions 1

See “Dual Attenuator Configuration:” on page 3

.

See “Single Attenuator Configuration:” on page 3

.

Key Path

Scope

Dependencies

Readback Line

Initial S/W Revision

Modified at S/W Revision

Dual Attenuator Configuration:

AMPTD Y Scale

Meas Global

In measurements which support the I/Q inputs, this key is unavailable when

I/Q is the selected input, and is replaced by the Range key in that case.

Contains a summary in [ ] brackets of the current total attenuation. See the descriptions of the

“(Mech) Atten ” on page 4 ,

“Enable Elec Atten” on page 6

, and

“Elec Atten” on page 8 keys for more detail on the contributors to the

total attenuation.

Note that when "Pre-Adjust for Min Clip" is on, this value can change at the start of every measurement.

Prior to A.02.00

A.03.00

20 Hz – 26.5 GHz

Input

2 2 6 10 20 30

High Band

(>~3.6 GHz)

Mech Attenuator

0-70 dB, 2 dB steps with switched DC block

Band

Switch

Limiter

25 dB Electronic

Attenuator

(optional)

Low Band

(<~3.6 GHz)

Single Attenuator Configuration:

9 kHz – 7.5 GHz

Input

20 dB

Fixed 20 dB stage Limiter 31 dB electronic stage

You can tell which attenuator configuration you have by pressing the Attenuation key, which (in most

Modes) opens up the Attenuation menu. If the first key in the Attenuation menu says

Mech Atten

you have the dual attenuator configuration. If the first key says

Atten

you have the single attenuator configuration.

Dual Attenuator Single Attenuator

3

Common Measurement Functions 1

In the single attenuator configuration, youcontrol the attenuation with a single control, as the fixed stage has only two states. In the dual attenuator configuration, both stages have significant range so you are given separate control of the mechanical and electronic attenuator stages.

When you have the dual attenuator configuration, you may still have only a single attenuator, because unless you purchase the Electronic Attenuator option you will only have the mechanical attenuator.

Most Attenuation settings are the same for all measurements – they do not change as you change measurements.

(Mech) Atten

This key is labeled

Mech Atten

in dual attenuator models and

Atten

in single attenuator models. In the dual attenuator configuration, this key only affects the mechanical attenuator.

This key lets you modify the attenuation applied to the RF input signal path. This value is normally auto coupled to the Ref Level, the Internal Preamp Gain, any External Gain that is entered, and the Max

Mixer Level, as described in the table below.

See

“Attenuator Configurations and Auto/Man” on page 5

.

Key Path

Remote Command

Example

Dependencies

AMPTD Y Scale, Attenuation

[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>

[:SENSe]:POWer[:RF]:ATTenuation?

[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1

[:SENSe]:POWer[:RF]:ATTenuation:AUTO?

POW:ATT 20

Dual attenuator configuration: sets the mechanical attenuator to 20 dB

Single attenuator mode: sets the main attenuation to 20 dB (see below for definition of “main” attenuation).

If the attenuator was in Auto, it sets it to Manual.

Some measurements do not support the Auto setting of

(Mech) Atten

. In these measurements, the Auto/Man selection is not available, and the

Auto/Man line on the key disappears.

In dual attenuator configurations, when the electronic attenuator is enabled, the mechanical attenuator has no auto setting and the Auto/Man line on the key disappears. The state of Auto/Man is remembered and restored when the electronic attenuator is once again disabled. This is described in more detail

in the

“Enable Elec Atten” on page 6

key description.

See “Attenuator Configurations and Auto/Man” on page 5

for more information on the Auto/Man functionality of Attenuation.

4

Common Measurement Functions 1

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

When

(Mech) Atten

is in Auto, it uses the following algorithm to determine a value:

Atten = ReferenceLevel + PreAmpGain + ExternalGain – RefLevelOffset -

MaxMixerLevel + IF Gain.

Limit this value to be between 6 dB and the Max value. No value below 6 dB can ever be chosen by Auto.

The resulting value is rounded up to the largest value possible given the attenuation step setting. That is, 50.01 dB would change to 60 dB (for a 10 dB attenuation step).

The “IF Gain” term in the equation above is either 0 dB or +10 dB, depending on the settings of FFT IF Gain, Swept IF Gain, max Ref Level and the

Auto/Man setting of Mech Atten.

Auto

Saved in instrument state

0 dB

The attenuation set by this key cannot be decreased below 6 dB with the knob or step keys. To get to a value below 6 dB it has to be directly entered from the keypad or via SCPI. This protects from adjusting the attenuation to a dangerously small value which can put the instrument at risk of damage to input circuitry. However, if the current mechanical attenuation is below 6 dB it can be increased with the knob and step keys, but not decreased.

CXA: 50 dB

EXA: 60 dB

MXA and PXA: 70 dB

In the single attenuator configuration, the total of ATT and EATT cannot exceed 50 dB, so if the EATT is set to 24 dB first, the main attenuation cannot be greater than 26 dB and will be reduced accordingly; if the main attenuator is set to 40 dB first, EATT cannot be greater than 10 dB.

Prior to A.02.00

A.03.00

Attenuator Configurations and Auto/Man

As described in the Attenuation key description, there are two distinct attenuator configurations available in the X-Series, the single attenuator and dual attenuator configurations. In dual attenuator configurations, we have the mechanical attenuation and the electronic attenuation, and the current total attenuation is the sum of the electronic + mechanical attenuation. In single attenuator configurations, we refer to the attenuation set using the

(Mech) Atten

key (or POW:ATT SCPI) as the “main” attenuation; and the attenuation that is set by the SCPI command POW:EATT as the “soft” attenuation (the

POW:EATT command is honored even in the single attenuator configuration, for compatibility purposes). Then the current total attenuation is the sum of the main + soft attenuation. See the

Elec

Atten

key description for more on “soft” attenuation.

In the dual attenuator configuration, when the electronic attenuator is enabled, there is no Auto/Man

5

Common Measurement Functions 1

functionality for the mechanical attenuator, and the third line of the key label (the Auto/Man line) disappears:

Mech Atten

0 dB

Auto Man

Mech Atten

0 dB

Mech Atten when elec atten disabled

Mech Atten when elec atten enabled vsd05

Enable Elec Atten

Enables the Electronic Attenuator in dual attenuator configurations. This key does not appear in single attenuator configurations, as in the single attenuator configuration there is no “electronic attenuator” there is only a single integrated attenuator (which has both a mechanical and electronic stage).

The electronic attenuator offers finer steps than the mechanical attenuator, has no acoustical noise, is faster, and is less subject to wear. These advantages primarily aid in remote operation and are negligible for front panel use. See

“Using the Electronic Attenuator: Pros and Cons” on page 8

for a detailed discussion of the pros and cons of using the electronic attenuator.

For the single attenuator configuration, for SCPI backwards compatibility, the “soft” attenuation feature replaces the dual attenuator configuration’s electronic attenuator. All the same couplings and limitations apply. See

“Attenuator Configurations and Auto/Man” on page 5

See

“More Information” on page 7

Remote Command

Example

Initial S/W Revision

Key Path

Dependencies

[:SENSe]:POWer[:RF]:EATTenuation:STATe OFF|ON|0|1

[:SENSe]:POWer[:RF]:EATTenuation:STATe?

POW:EATT:STAT ON

Prior to A.02.00

AMPTD Y Scale, Attenuation

This key only appears in the dual attenuator configuration. However, in the single attenuator configuration, EATT SCPI commands are accepted for compatibility with other X-series instruments and set a “soft” attenuation as described in

“Attenuator Configurations and Auto/Man” on page 5

The electronic attenuator (and the “soft” attenuation function provided in single attenuator configurations) is unavailable above 3.6 GHz. Therefore, if the Stop Frequency of the analyzer is > 3.6 GHz then the

Enable Elec Atten

key will be OFF and grayed out.

If the Internal Preamp is on, meaning it is set to Low Band or Full, the electronic attenuator (and the “soft” attenuation function provided in single attenuator configurations) is unavailable. In this case the

Enable Elec Atten

key will be OFF and grayed out.

If either of the above is true, if the SCPI command is sent, an error indicating that the electronic attenuator is unavailable will be sent.

If the electronic/soft Attenuator is enabled, then the Stop Freq of the analyzer is limited to 3.6 GHz and the Internal Preamp is unavailable.

6

Common Measurement Functions 1

Couplings

Preset

State Saved

Modified at S/W Revision

Enabling and disabling the Electronic Attenuator affects the setting of the

Mechanical Attenuator (in dual attenuator configurations). This is described in more detail below this table.

OFF for Swept SA measurement; ON for all other measurements that support the electronic attenuator

Saved in instrument state

A.03.00

More Information

When the Electronic Attenuator is enabled, the Mechanical Attenuator transitions to a state in which it has no Auto function. Here are the rules for transitioning the Mechanical Attenuator:

When the Electronic Attenuation is enabled:

• In the dual attenuator configuration, the Mechanical Attenuator is initialized to 10 dB (this is its optimal performance setting). You can then set it as desired with SCPI, numeric keypad, step keys, or knob, and it behaves as it normally would in manual mode

• The Auto/Man state of (Mech) Atten is saved

• The Auto/Man line on the (Mech) Atten key disappears and the auto rules are disabled

• In the dual attenuator configuration, the Electronic Attenuator is set to 10 dB less than the previous value of the Mechanical Attenuator, within the limitation that it must stay within the range of 0 to 24 dB of attenuation.

Examples in the dual attenuator configuration:

• Mech Atten at 20 dB. Elec Atten enabled, Mech Atten set to 10 dB, and Elec Atten set to 10 dB. New total attenuation equals the value before Elec Atten enabled.

• Mech Atten at 0 dB. Elec Atten enabled, Mech Atten set to 10 dB, and Elec Atten set to 0 dB. New total attenuation does not equal the value before Elec Atten enabled.

• Mech Atten at 40 dB. Elec Atten enabled, Mech Atten set to 10 dB, and Elec Atten set to 24 dB. New total attenuation does not equal the value before Elec Atten enabled.

When the Electronic Attenuation is disabled:

• In the dual attenuator configuration, the Elec Atten key is grayed out (it never displays in the single attenuator configuration)

• The Auto/Man state of (Mech) Atten is restored

• If now in Auto, (Mech) Atten recouples

• If now in Man, (Mech) Atten is set to the value of total attenuation that existed before the Elec Atten was disabled. The resulting value is rounded up to the smallest value possible given the (Mech) Atten

Step setting - (That is, 57 dB changes to 58 dB when (Mech) Atten Step is 2 dB.)

7

Common Measurement Functions 1

Using the Electronic Attenuator: Pros and Cons

The electronic attenuator offers finer steps than the mechanical attenuator, has no acoustical noise, is faster, and is less subject to wear.

The “finer steps” advantage of the electronic attenuator is beneficial in optimizing the alignment of the analyzer dynamic range to the signal power in the front panel as well as remote use. Thus, you can achieve improved relative signal measurement accuracy. Compared to a mechanical attenuator with 2 dB steps, the 1 dB resolution of the electronic attenuator only gives better resolution when the odd-decibel steps are used. Those odd-decibel steps are less accurately calibrated than the even-decibel steps, so one tradeoff for this superior relative accuracy is reduced absolute amplitude accuracy.

Another disadvantage of the electronic attenuator is that the spectrum analyzer loses its “Auto” setting, making operation less convenient.

Also, the relationship between the dynamic range specifications (TOI, SHI, compression and noise) and instrument performance are less well-known with the electrical attenuator. With the mechanical attenuator, TOI, SHI and compression threshold levels increase dB-for-dB with increasing attenuation, and the noise floor does as well. With the electronic attenuator, there is an excess attenuation of about 1 to 3 dB between 0 and 3.6 GHz, making the effective TOI, SHI, and so forth, less well known. Excess attenuation is the actual attenuation relative to stated attenuation. Excess attenuation is accounted for in the analyzer calibration

Elec Atten

Controls the Electronic Attenuator in dual attenuator configurations. This key does not appear in single attenuator configurations, as the control of both the mechanical and electronic stages of the single attenuator is integrated into the single

Atten

key.

Key Path

Remote Command

Notes

Dependencies

Preset

State Saved

Min

AMPTD Y Scale, Attenuation

[:SENSe]:POWer[:RF]:EATTenuation <rel_ampl>

[:SENSe]:POWer[:RF]:EATTenuation?

Electronic Attenuation’s spec is defined only when Mechanical Attenuation is

6 dB.

This key only appears in the dual attenuator configuration. However, in the single attenuator configuration, EATT SCPI commands are accepted for compatibility with other X-series instruments and set a “soft” attenuation as described in

“Attenuator Configurations and Auto/Man” on page 5

. The

“soft” attenuation is treated as an addition to the “main” attenuation value set by the Atten softkey or the POW:ATT SCPI command and affects the total attenuation displayed on the Attenuation key and the Meas Bar.

When

Enable Elec Atten

is off, the

Elec Atten

key is grayed out.

0 dB

Saved in instrument state

0 dB

8

Common Measurement Functions 1

Max Dual attenuator configuration: 24 dB

Single attenuator configuration: the total of ATT and EATT cannot exceed 50 dB, so if the EATT is set to 24 dB first, the main attenuation cannot be greater than 26 dB and will be reduced accordingly; if the main attenuator is set to 40 dB first, EATT cannot be greater than 10 dB

Prior to A.02.00

A.03.00

Initial S/W Revision

Modified at S/W Revision

Adjust Atten for Min Clip

Sets the combination of mechanical and electronic attenuation based on the current measured signal level so that clipping will be at a minimum.

This is an "immediate action" function, that is, it executes once, when the key is pressed.

This key is grayed out in measurements that do not support this functionality. The spectrum analyzer measurement, Swept SA, does not support this functionality.

Key Path

Remote Command

Initial S/W Revision

Modified at S/W Revision

AMPTD Y Scale, Attenuation

[:SENSe]:POWer[:RF]:RANGe:OPTimize IMMediate

Prior to A.02.00

A.03.00

Pre-Adjust for Min Clip

If this function is on, it does the adjustment described under “Adjust Atten for Min Clip” on page 9

each time a measurement restarts. Therefore, in Continuous measurement mode, it only executes before the first measurement.

In dual attenuator models, you can set

Elec+Mech Atten

, in which case both attenuators participate in the autoranging, or

Elec Atten Only

, in which case the mechanical attenuator does not participate in the autoranging. This latter case results in less wear on the mechanical attenuator and is usually faster.

This key is grayed out in measurements that do not support this functionality. The spectrum analyzer measurement, Swept SA, does not support this functionality.

Remote Command

Initial S/W Revision

Key Path

Notes

[:SENSe]:POWer[:RF]:RANGe:OPTimize:ATTenuation

OFF|ELECtrical|COMBined

[:SENSe]:POWer[:RF]:RANGe:OPTimize:ATTenuation?

Prior to A.02.00

AMPTD Y Scale, Attenuation

The SCPI parameter ELECtrical sets this function to On in single attenuator models. The SCPI parameter COMBined is mapped to ELECtrical in single attenuator models; if you send COMBined, it sets the function to On and returns ELEC to a query.

9

Common Measurement Functions 1

Preset

State Saved

Range

Modified at S/W Revision

OFF for Swept SA measurement; ON for all other measurements that support

Pre-Adjust for Min Clip

Saved in instrument state

Dual attenuator models:

Off | Elec Atten Only | Mech + Elec Atten

Single attenuator models:

Off | On

A.03.00

Remote Command:

Notes:

[:SENSe]:POWer[:RF]:RANGe:AUTO ON|OFF|1|0

[:SENSe]:POWer[:RF]:RANGe:AUTO?

ON aliases to "Elec Atten Only"

OFF aliases to "Off"

The query returns true if not "Off"

Prior to A.02.00

Initial S/W Revision:

(Mech) Atten Step

This controls thestep size used when making adjustments to the input attenuation.

This key is labeled

Mech Atten Step

in dual attenuator models and

Atten Step

in single attenuator models. In the dual attenuator configuration, this key only affects the step size of the mechanical attenuator.

Key Path

Remote Command

Example

Notes

Dependencies

Couplings

Preset

State Saved

AMPTD Y Scale, Attenuation

[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement] 10 dB |

2 dB

[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement]?

POW:ATT:STEP 2

Note this feature works like a 1-N choice from the front panel, but it takes a specific value (in dB) when used remotely. The only valid values are 2 and 10.

Blanked in CXA and EXA if option FSA (2 dB steps) is not present. If blanked, attempts to set it via SCPI will yield an "Option not present" error.

When the attenuation step size changes, the current mechanical attenuation value is adjusted (if necessary) to be quantized to the new step size. That is, if step is set to 10 dB, mech atten is increased if necessary so it is a multiple of

10 dB

PXA and MXA: 2 dB

EXA and CXA: 10 dB (2 dB with option FSA)

Saved in instrument state

10

Common Measurement Functions 1

Initial S/W Revision

Modified at S/W Revision

Prior to A.02.00

A.03.00

Max Mixer Level

Controls the limitation on the Ref Level for a given attenuation setting, and therefore also interacts with the Auto rules for selecting the attenuation as a coupling from the reference level.

Key Path

Remote Command

Example

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

AMPTD Y Scale, Attenuation

[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer] <real>

[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?

POW:MIX:RANG –15 dBm

–10 dBm

Saved in instrument state

–50 dBm

–10 dBm

Depends on the current selected Y axis unit, see Swept SA discussion of Y

Axis Unit

Prior to A.02.00

Range

This key is only available when I/Q is the selected input. It replaces the Attenuation key in that case.

Each input channel (I and Q) has four internal gain ranges. The maximum allowed voltage in each gain range is slightly more than the nominal value, so the break point between ranges is a couple of millivolts higher than the nominal (setting a peak voltage of 0.502 mV will still map to the 0.5 V Peak range).

Gain Setting

0 dB

6 dB

12 dB

18 dB

Volts RMS

0.7071

0.3536

0.1768

0.0884

Volts Peak

1.0

0.5

0.25

0.125

Volts Peak - Peak

2.0

1.0

0.5

0.25

dBm (50

Ω)

10

4

–2

–8

Break Point n/a

0.502 V Peak

0.252 V Peak

0.127 V Peak

Key Path

Notes

State Saved

AMPTD Y Scale

Visible only when the selected input is I/Q.

No

11

Common Measurement Functions 1

Readback Text When Range is Auto, "[Auto]"

When Range is Man and I & Q are the same, "[<range value>]"

When Range is Man and I & Q are different:

"[I: <I range value>

Q: <Q range value>]"

See I Range and Q Range for the <range value> enumeration definition.

Prior to A.02.00

Initial S/W Revision

Range Auto/Man

The Auto setting for Range causes the range to be set based on the Y Scale settings. When Range is

“Auto”, the I & Q Range are set based on the top of the Y Scale when the Y scale is in dB units (for example, power), or to the max(abs(top), abs(bottom)) when the Y scale reference is not at the top of the screen.

Not all measurements support Range Auto/Man. If Auto is not supported in the current measurement, this key is grayed out and shows “Man” and MAN is returned to a SCPI query; but this does NOT change the Auto/Man setting for Range.When you go to a measurement that supports Auto, it goes back to Auto if it was previously in Auto mode.

Key Path

Scope

Remote Command

Example

Dependencies

Couplings

Preset

State Saved

Range

Initial S/W Revision

AMPTD Y Scale, Range

Meas Global

[:SENSe]:VOLTage:IQ:RANGe:AUTO OFF|ON|0|1

[:SENSe]:VOLTage:IQ:RANGe:AUTO?

Put the I Range and Q Range in manual.

VOLT:IQ:RANG:AUTO OFF

If Auto is not supported, sending the SCPI command will generate an error.

When in Auto, both I Range and Q Range are set to the same value, computed as follows:

Maximum absolute value is computed for the Y Scale. The top and bottom of the graph are computed based on Ref Value, Scale/Div, and Ref Position.

Formula: YMax = max(abs(top), abs(bottom)).

The I Range and Q Range are then set to YMax.

ON

Saved in instrument state

Auto | Man

Prior to A.02.00

Remote Command:

[:SENSe]:POWer:IQ:RANGe:AUTO OFF|ON|0|1

[:SENSe]:POWer:IQ:RANGe:AUTO?

12

Common Measurement Functions 1

Example:

Notes:

Preset:

Range:

Initial S/W Revision:

Put the I Range and Q Range in manual.

POW:IQ:RANG:AUTO OFF

The POW:IQ:RANG:AUTO is an alternate form of the

VOLT:IQ:RANG:AUTO command. This is to maintain consistency with I

Range and Q Range, which support both the POWer and VOLTage forms of the command.

ON

Auto | Man

Prior to A.02.00

I Range

This is the internal gain range for the I channel when Input Path is I Only or Ind I/Q, and it is used for

both the I and Q channels when the Input Path is I+jQ. See “I/Q Gain Ranges” on page 16

.

Key Path

Remote Command

Example

Notes

Couplings

Preset

State Saved

Range

Initial S/W Revision

AMPTD Y Scale, Range

[:SENSe]:VOLTage:IQ[:I]:RANGe[:UPPer] <voltage>

[:SENSe]:VOLTage:IQ[:I]:RANGe[:UPPer]?

Set the I Range to 0.5 V Peak

VOLT:IQ:RANG 0.5 V

The numeric entries are mapped to the smallest gain range whose break point is greater than or equal to the value, or 1 V Peak if the value is greater than 1

V.

When Q Same as I is On, the I Range value will be copied to the Q Range.

Changing the value will also set Range = Man.

1 V Peak

Saved in instrument state

1 V Peak | 0.5 V Peak | 0.25 V Peak | 0.125 V Peak

Prior to A.02.00

Remote Command:

Example:

[:SENSe]:POWer:IQ[:I]:RANGe[:UPPer] <ampl>

[:SENSe]:POWer:IQ[:I]:RANGe[:UPPer]?

Set the I Range to 0.5 V Peak when Reference Z is 50

Ω, and to 1.0 V Peak when Reference Z is 75

Ω.

POW:IQ:RANG 4 dBm

13

Common Measurement Functions 1

Notes:

Preset:

Range:

Min:

Max:

Initial S/W Revision:

Q Range

Accesses the Q Range menu.

Key Path

Readback Text

The POWer form of the command is provided for convenience. It maps to the same underlying gain range parameter as the VOLTage form of the command.

The Reference Z (not the I channel Input Z) is used to convert the power to peak voltage, which is then used to set the I Range as with the VOLTage form of the command. The power values of the 4 range states (1V Peak, 0.5V Peak,

0.25V Peak, and 0.125V Peak) will vary with Reference Z. Here are some examples:

50

Ω: 10, 4, –2, –8

75

Ω: 8.2, 2.2, –3.8, –9.8

600

Ω: –0.8, –6.8, –12.8, –18.9

10.0 dBm

–20 dBm to 10 dBm

–20 dBm

10 dBm

Prior to A.02.00

AMPTD Y Scale, Range

Q Same as I | 1 V Peak | 0.5 V Peak | 0.25 V Peak | 0.125 V Peak

When Q Same as I is On, the readback is "Q Same as I", otherwise it is the Q

Range value.

Prior to A.02.00

Initial S/W Revision

Q Same as I

Many, but not all, usages require the I and Q channels to have an identical setup. To simplify channel setup, the Q

Same as I will cause the Q channel range to be mirrored from the I channel. That way you only need to set up one channel (the I channel). The I channel values are copied to the Q channel, so at the time Q Same as I is Off, the I and Q channel setups will be identical.

Key Path

Remote Command

Example

Couplings

Preset

AMPTD Y Scale, Range, Q Range

[:SENSe]:VOLTage|POWer:IQ:MIRRored OFF|ON|0|1

[:SENSe]:VOLTage|POWer:IQ:MIRRored?

Turn off the mirroring of I Range to Q Range.

VOLT:IQ:MIRR OFF

POW:IQ:MIRR OFF

When On, the I Range value is mirrored (copied) to the Q Range.

On

14

Common Measurement Functions 1

State Saved

Range

Readback Text

Initial S/W Revision

Saved in instrument state.

On | Off

"Q Same as I" when On, otherwise none.

Prior to A.02.00

Q Range Value

This is the internal gain range for the Q channel. See “I/Q Gain Ranges” on page 16 . The Q Range only applies to

Input Path Q Only and Ind I/Q. For input I+jQ the I Range determines both I and Q channel range settings.

Key Path

Remote Command

Example

Notes

Couplings

Preset

State Saved

Range

Initial S/W Revision

AMPTD Y Scale, Range

[:SENSe]:VOLTage:IQ:Q:RANGe[:UPPer] <voltage>

[:SENSe]:VOLTage:IQ:Q:RANGe[:UPPer]?

Set the Q Range to 0.5 V Peak

VOLT:IQ:Q:RANG 0.5 V

The numeric entries are mapped to the smallest gain range whose break point is greater than or equal to the value, or 1 V Peak if the value is greater than 1

V.

The Q Range is only used for Input Path Q Only and Ind I/Q. For input I+jQ the I Range determines both I and Q channel range settings.

When Q Same as I is On, the I Range value will be copied to the Q Range and the range value keys are disabled.

Changing the value will also set Range = Man.

1 V Peak

Saved in instrument state

1 V Peak | 0.5 V Peak | 0.25 V Peak | 0.125 V Peak

Prior to A.02.00

Remote Command:

Example:

[:SENSe]:POWer:IQ:Q:RANGe[:UPPer] <ampl>

[:SENSe]:POWer:IQ:Q:RANGe[:UPPer]?

Will set the Q Range to 0.5 V Peak when Reference Z is 50

Ω, and to 1.0 V

Peak when Reference Z is 75

Ω.

POW:IQ:Q:RANG 4 dBm

15

Common Measurement Functions 1

Notes:

Preset:

Range:

Min:

Max:

Initial S/W Revision:

I/Q Gain Ranges

1 V Peak

Set the channel gain state to 1 Volt Peak.

The POWer form of the command is provided for convenience. It maps to the same underlying gain range parameter as the VOLTage form of the command.

The Reference Z (not the Q channel Input Z) is used to convert the power to peak voltage, which is then used to set the Q Range as with the VOLTage form of the command. The power values of the 4 range states (1V Peak, 0.5V Peak,

0.25V Peak, and 0.125V Peak) will vary with Reference Z. Here are some examples:

50

Ω: 10, 4, –2, –8

75

Ω: 8.2, 2.2, –3.8, –9.8

600

Ω: –0.8, –6.8, –12.8, –18.9

10.0 dBm

–20 dBm to 10 dBm

–20 dBm

10 dBm

Prior to A.02.00

Key Path

Initial S/W Revision

AMPTD Y Scale, I Range | Q Range

Prior to A.02.00

0.5 V Peak

Set the channel gain state to 0.5 Volt Peak.

Key Path

Initial S/W Revision

AMPTD Y Scale, I Range | Q Range

Prior to A.02.00

0.25 V Peak

Set the channel gain state to 0.25 Volt Peak.

Key Path

Initial S/W Revision

AMPTD Y Scale, I Range | Q Range

Prior to A.02.00

0.125 V Peak

Set the channel gain state to 0.125 Volt Peak.

Key Path

AMPTD Y Scale, I Range | Q Range

16

NOTE

Common Measurement Functions 1

Initial S/W Revision Prior to A.02.00

Scale / Div

Sets the units per vertical graticule division on the display. This function is only available when Scale

Type (Log) is selected and the vertical scale is power. When Scale Type (Lin) is selected, Scale/Div is grayed out.

Key Path

Remote Command

Example

Dependencies

Preset

State Saved

Min

Max

Initial S/W Revision

AMPTD Y Scale

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:PDIVision <rel_ampl>

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:PDIVision?

DISP:WIND:TRAC:Y:PDIV 5 DB

Scale/Div is grayed out in linear Y scale. Sending the equivalent SCPI command does change the Scale/Div, though it has no affect while in Lin.

10.00 dB / Div

Saved in instrument state

0.10 dB

20 dB

Prior to A.02.00

Scale Type

Chooses a linear or logarithmic vertical scale for the display and for remote data readout.

When Scale Type (Log) is selected, the vertical graticule divisions are scaled in logarithmic units. The top line of the graticule is the Reference Level and uses the scaling per division Scale/Div to assign values to the other locations on the graticule.

When Scale Type (Lin) is selected, the vertical graticule divisions are linearly scaled with the reference level value at the top of the display and zero volts at the bottom. Each vertical division of the graticule represents one-tenth of the Reference Level.

The Y Axis Unit used for each type of display is set by pressing Y Axis Unit. The analyzer remembers separate Y Axis Unit settings for both Log and Lin.

Key Path

Remote Command

AMPTD Y Scale

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:SPACing

LINear|LOGarithmic

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:SPACing?

17

Common Measurement Functions 1

Example

Dependencies

Couplings

Preset

State Saved

Initial S/W Revision

DISP:WIND:TRAC:Y:SPAC LOG

DISP:WIND:TRAC:Y:SPAC?

If Normalize is on, Scale Type forced to Log and is grayed out.

Changing the Scale Type always sets the Y Axis unit to the last unit specified for the current amplitude scale. In other words, we restore the Y Axis unit setting appropriate per log/lin.

LOG

Saved in instrument state

Prior to A.02.00

Presel Center

When this key is pressed, the centering of the preselector filter is adjusted to optimize the amplitude accuracy at the frequency of the selected marker. If the selected marker is not on when Presel Center is pressed, the analyzer will turn on the selected marker, perform a peak search, and then perform centering on the marker’s center frequency. If the selected marker is already on and between the start and stop frequencies of the analyzer, the analyzer performs the preselector calibration on that marker’s frequency.

If the selected marker is already on, but outside the frequency range between Start Freq and Stop Freq, the analyzer will first perform a peak search, and then perform centering on the marker’s center frequency.

The value displayed on the

Presel Adjust

key will change to reflect the new preselector tuning (see

Presel Adjust

.

A number of considerations should be observed to ensure proper operation. See

“Proper Preselector

Operation” on page 19 .

Key Path

Remote Command

Example

Notes

Dependencies

AMPTD Y Scale

[:SENSe]:POWer[:RF]:PCENter

POW:PCEN

Note that the rules outlined above under the key description apply for the remote command as well as the key. The result of the command is dependent on marker position, and so forth. Any message shown by the key press is also shown in response to the remote command.

• Grayed out if the microwave preselector is off. )

• If the selected marker’s frequency is below Band 1, advisory 0.5001 is generated and no action is taken.

• Grayed out if entirely in Band 0.

• Blank in models that do not include a preselector, such as option 503. If the SCPI is sent in these instruments, it is accepted without error, and the query always returns 0.

18

Common Measurement Functions 1

Couplings

Status Bits/OPC dependencies

The active marker position determines where the centering will be attempted.

If the analyzer s in a measurement such as averaging when centering is initiated, the act of centering the preselector will restart averaging but the first average trace will not be taken until the centering is completed.

When centering the preselector, *OPC will not return true until the process is complete and a subsequent measurement has completed, nor will results be returned to a READ or MEASure command.

The Measuring bit should remain set while this command is operating and should not go false until the subsequent sweep/measurement has completed.

Prior to A.02.00

A.03.00

Initial S/W Revision

Modified at S/W Revision

Proper Preselector Operation

A number of considerations should be observed to ensure proper operation:

If the selected marker is off, the analyzer will turn on a marker, perform a peak search, and adjust the preselector using the selected marker’s frequency. It uses the "highest peak" peak search method unqualified by threshold or excursion, so that there is no chance of a ‘no peak found’ error. It continues with that peak, even if it is the peak of just noise. Therefore, for this operation to work properly, there should be a signal on screen in a preselected range for the peak search to find.

If the selected marker is already on, the analyzer will attempt the centering at that marker’s frequency.

There is no preselector for signals below about 3.6 GHz, therefore if the marker is on a signal below 3.6

GHz, no centering will be attempted and an advisory message generated

In some models, the preselector can be bypassed. If it is bypassed, no centering will be attempted in that range.

Preselector Adjust

Allows you to manually adjust the preselector filter frequency to optimize its response to the signal of interest. This function is only available when

“Presel Center” on page 18

is available.

For general purpose signal analysis, using Presel Center is recommended. Centering the filter minimizes the impact of long-term preselector drift. Presel Adjust can be used instead to manually optimize the preselector. One application of manual optimization would be to peak the preselector response, which both optimizes the signal-to-noise ratio and minimizes amplitude variations due to small (short-term) preselector drifting.

Key Path

Scope

Remote Command

Example

AMPTD Y Scale

Meas Global

[:SENSe]:POWer[:RF]:PADJust <freq>

[:SENSe]:POWer[:RF]:PADJust?

POW:PADJ 100KHz

POW:PADJ?

19

Common Measurement Functions 1

Notes

Dependencies

Preset

State Saved

Min

Max

Default Unit

Backwards Compatibility SCPI

Initial S/W Revision

Modified at S/W Revision

The value on the key reads out to 0.1 MHz resolution.

• Grayed out if microwave preselector is off. )

• Grayed out if entirely in Band 0.

• Blank in models that do not include a preselector, such as option 503. If the SCPI is sent in these instruments, it is accepted without error, and the query always returns 0.

0 MHz

The

Presel Adjust

value set by

Presel Center

, or by manually adjusting

Presel Adjust

, is not saved in Instrument State, and does not survive a Preset or power cycle.

–500 MHz

500 MHz

Hz

[:SENSe]:POWer[:RF]:MW:PADJust

[:SENSe]:POWer[:RF]:MMW:PADJust

(These were undocumented commands for PSA which X-Series will accept)

Prior to A.02.00

A.03.00

Remote Command:

Notes:

[:SENSe]:POWer[:RF]:PADJust:PRESelector

MWAVe|MMWave|EXTernal

[:SENSe]:POWer[:RF]:PADJust:PRESelector?

[:SENSe]:POWer[:RF]:PADJust:PRESelector MWAVe|MMWave|EXTernal where: MWAV = 3–26 GHz

MMWave = 26–50 GHz

EXTernal = External

Preselector Selection - PSA had multiple preselectors, and you could select which preselector to center. Since the X-Series will have only one preselector, the preselector selection softkey will no longer be available. However, in order to provide backward compatibility, we will support the remote command.

The command form is a NOP

The query will return MWAVe

Prior to A.02.00

Initial S/W Revision:

Y Axis Unit

Displays the menu keys that enable you to change the vertical (Y) axis amplitude unit. The analyzer retains the entered Y Axis Unit separately for both Log and Lin amplitude scale types. For example, if

20

Common Measurement Functions 1

Scale Type has been set to Log, and you set Y Axis Unit to dBm, pressing Scale Type (Log) sets the Y

Axis Unit to dBm. If Scale Type has been set to Lin and you set Y Axis Unit to V, pressing Scale Type

(Lin) sets the Y Axis Unit to V. Pressing Scale Type (Log) again sets the Y axis unit back to dBm.

NOTE

The units of current (A, dBmA, dBuA) are calculated based on 50 ohms input impedance.

All four of the EMI units (dB

μA/m, dBμV/m, dBG, dBpT) are treated by the instrument exactly as though they were dBuV. The user must load an appropriate correction factor using Amplitude Corrections for accurate and meaningful results.

If a SCPI command is sent to the analyzer which uses one of the EMI units as a terminator, the analyzer treats it as though DBUV had been sent as the terminator.

Remote Command

Example

Initial S/W Revision

Key Path

Scope

Notes

:UNIT:POWer

DBM|DBMV|DBMA|V|W|A|DBUV|DBUA|DBUVM|DBUAM|DBPT|DBG

:UNIT:POWer?

UNIT:POW dBmV

UNIT:POW?

Prior to A.02.00

AMPTD Y Scale

Meas Global

The Y axis unit has either logarithmic or linear characteristics. The set of units that is logarithmic consists of dBm, dBmV, dBmA, dB

μV, dBμA, dBμV/m, dB

μA/m, dBpT, and dBG. The set if units that is linear consists of V, W, and

A. The chosen unit will determine how the reference level and all the amplitude-related outputs like trace data, marker data, etc. read out.

21

Common Measurement Functions 1

Notes

Dependencies

Couplings

Preset

State Saved

Readback line

Modified at S/W Revision

The settings of Y Axis Unit and Scale Type, affect how the data is read over the remote interface. When using the remote interface no unit is returned, so you must know what the Y axis unit is to interpret the results:

Example 1, set the following:

Scale Type (Log)

Y Axis Unit, dBm

Scale/Div, 1 dB

Ref Level, 10 dBm

This sets the top line to 10 dBm with each vertical division representing 1 dB.

Thus, if a point on trace 1 is on the fifth graticule line from the top, it represents 5 dBm and will read out remotely as 5.

Example 2, set the following:

Scale Type (Lin)

Y Axis Unit, Volts

Ref Level, 100 mV (10 mV/div)

This sets the top line to 100 mV and the bottom line to 0 V, so each vertical division represents 10 mV. Thus, if a point on trace 1 is on the fifth graticule line from the top, it represents 50 mV and will read out remotely as 50.

If an amplitude correction with an Antenna Unit other than

None

is applied and enabled, then that antenna unit is forced and the key with that unit is the only Y Axis Unit available. All other Y Axis Unit keys are grayed out.

If an amplitude correction with an Antenna Unit other than

None

is applied and enabled, and you then turn off that correction or set

Apply Corrections

to

No

, the Y Axis Unit that existed before the Antenna Unit was applied is restored.

The analyzer retains the entered Y Axis Unit separately for both Log and Lin amplitude scale types dBm for log scale, V for linear. The true ‘preset’ value is dBm, since at preset the Y Scale type is set to logarithmic.

Saved in instrument state

1-of-N selection

A.02.00, A.04.00

dBm

Sets the amplitude unit for the selected amplitude scale (log/lin) to dBm.

Key Path

Example

Dependencies

AMPTD Y Scale, Y Axis Unit

UNIT:POW DBM

Grayed out if an Amplitude Correction with an Antenna Unit is ON.

22

Common Measurement Functions 1

Readback

Initial S/W Revision

Modified at S/W Revision dBm

Prior to A.02.00

A.02.00

dBmV

Sets the amplitude unit for the selected amplitude scale (log/lin) to dBmV.

Key Path

Example

Dependencies

Readback

Initial S/W Revision

Modified at S/W Revision

AMPTD Y Scale, Y Axis Unit

UNIT:POW DBMV

Grayed out if an Amplitude Correction with an Antenna Unit is ON. dBmV

Prior to A.02.00

A.02.00

dBmA

Sets the amplitude unit for the selected amplitude scale (log/lin) to dBmA.

Key Path

Example

Dependencies

Readback

Initial S/W Revision

Modified at S/W Revision

AMPTD Y Scale, Y Axis Unit

UNIT:POW DBMA

Grayed out if an Amplitude Correction with an Antenna Unit is ON. dBmA

Prior to A.02.00

A.02.00

W

Sets the amplitude unit for the selected amplitude scale (log/lin) to watt.

Key Path

Example

Dependencies

Readback

Initial S/W Revision

Modified at S/W Revision

AMPTD Y Scale, Y Axis Unit

UNIT:POW W

Grayed out if an Amplitude Correction with an Antenna Unit is ON.

W

Prior to A.02.00

A.02.00

23

Common Measurement Functions 1

V

Sets the amplitude unit for the selected amplitude scale (log/lin) to volt.

Key Path

Example

Dependencies

Readback

Initial S/W Revision

Modified at S/W Revision

AMPTD Y Scale, Y Axis Unit

UNIT:POW V

Grayed out if an Amplitude Correction with an Antenna Unit is ON.

V

Prior to A.02.00

A.02.00

A

Sets the amplitude unit for the selected amplitude scale (log/lin) to Ampere.

Key Path

Example

Dependencies

Readback

Initial S/W Revision

Modified at S/W Revision

AMPTD Y Scale, Y Axis Unit

UNIT:POW A

Grayed out if an Amplitude Correction with an Antenna Unit is ON.

A

Prior to A.02.00

A.02.00

dB

μV

Sets the amplitude unit for the selected amplitude scale (log/lin) to dB

μV.

Key Path

Example

Dependencies

Readback

Initial S/W Revision

Modified at S/W Revision

AMPTD Y Scale, Y Axis Unit

UNIT:POW DBUV

Grayed out if an Amplitude Correction with an Antenna Unit is ON. dB

μV

Prior to A.02.00

A.02.00

dB

μA

Sets the amplitude unit for the selected amplitude scale (log/lin) to dB

μA.

Key Path

Example

Dependencies

AMPTD Y Scale, Y Axis Unit

UNIT:POW DBUA

Grayed out if an Amplitude Correction with an Antenna Unit is ON.

24

Common Measurement Functions 1

Readback

Initial S/W Revision

Modified at S/W Revision dB

μA

Prior to A.02.00

A.02.00

dB

μV/m

Sets the amplitude unit for the selected amplitude scale (log/lin) to dB

μV/m. This is an antenna unit, and this key is grayed out unless a Correction with this Antenna Unit selected is ON. If this is the case, all of the other Antenna Units are grayed out.

Key Path

Example

Dependencies

Readback

Initial S/W Revision

AMPTD Y Scale, Y Axis Unit

UNIT:POW DBUVM

Grayed out if no Amplitude Correction with an Antenna Unit is on. dB

μV/m

A.02.00

dB

μA/m

Sets the amplitude unit for the selected amplitude scale (log/lin) to dB

μA/m. This is an antenna unit, and this key is grayed out unless a Correction with this Antenna Unit selected is ON. If this is the case, all of the other Antenna Units are grayed out.

Key Path

Example

Dependencies

Readback

Initial S/W Revision

AMPTD Y Scale, Y Axis Unit

UNIT:POW DBUAM

Grayed out if no Amplitude Correction with an Antenna Unit is on. dB

μA/m

A.02.00

dBpT

Sets the amplitude unit for the selected amplitude scale (log/lin) to dBpT. This is an antenna unit, and this key is grayed out unless a Correction with this Antenna Unit selected is ON. If this is the case, all of the other Antenna Units are grayed out.

Key Path

Example

Dependencies

Readback

Initial S/W Revision

AMPTD Y Scale, Y Axis Unit

UNIT:POW DBPT

Grayed out if no Amplitude Correction with an Antenna Unit is on. dBpT

A.02.00

25

Common Measurement Functions 1

dBG

Sets the amplitude unit for the selected amplitude scale (log/lin) to dBG. This is an antenna unit, and this key is grayed out unless a Correction with this Antenna Unit selected is ON. If this is the case, all of the other Antenna Units are grayed out.

Key Path

Example

Dependencies

Readback

Initial S/W Revision

AMPTD Y Scale, Y Axis Unit

UNIT:POW DBG

Grayed out if no Amplitude Correction with an Antenna Unit is on. dBG

A.02.00

Reference Level Offset

Adds an offset value to the displayed reference level. The reference level is the absolute amplitude represented by the top graticule line on the display.

See

“More Information” on page 26

Remote Command

Example

Initial S/W Revision

Key Path

Scope

Preset

State Saved

Min

Max

Modified at S/W Revision

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel:OFFSet

<rel_ampl>

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel:OFFSet?

DISP:WIND:TRAC:Y:RLEV:OFFS 12.7

Sets the Ref Level Offset to 12.7 dB. The only valid suffix is dB. If no suffix is sent, dB will be assumed.

Prior to A.02.00

AMPTD Y Scale

Meas Global

0 dBm

Saved in instrument state

The range for Ref Lvl Offset is variable. It is limited to values that keep the reference level within the range of –327.6 dB to 327.6 dB.

327.6 dB

A.04.00

More Information

Offsets are used when gain or loss occurs between a device under test and the analyzer input. Thus, the signal level measured by the analyzer may be thought of as the level at the input of an external amplitude conversion device. Entering an offset does not affect the trace position or attenuation value, just the value of the top line of the display and the values represented by the trace data. Thus, the values of exported trace data, queried trace data, marker amplitudes, trace data used in calculations such as N dB points,

26

Common Measurement Functions 1

trace math, peak threshold, and so forth, are all affected by Ref Level Offset.

NOTE

Changing the offset causes the analyzer to immediately stop the current sweep and prepare to begin a new sweep, but the data will not change until the trace data updates, because the offset is applied to the data as it is taken. If a trace is exported with a nonzero Ref Level Offset, the exported data will contain the trace data with the offset applied.

The maximum reference level available is dependent on the reference level offset. That is, Ref Level -

Ref Level Offset must be in the range –170 to +30 dBm. For example, the reference level value range can be initially set to values from –170 dBm to 30 dBm with no reference level offset. If the reference level is first set to –20 dBm, then the reference level offset can be set to values of –150 to +50 dB.

If the reference level offset is first set to –30 dB, then the reference level can be set to values of –200 dBm to 0 dBm. In this case, the reference level is “clamped” at 0 dBm because the maximum limit of

+30 dBm is reached with a reference level setting of 0 dBm with an offset of –30 dB. If instead, the reference level offset is first set to 30 dB, then the reference level can be set to values of –140 to +60 dBm.

µW Path Control

The

µW Path Control

functions include the

µW Preselector Bypass

(Option MPB) and

Low Noise

Path

(Option LNP) controls in the High Band path circuits.

When the µW Preselector is bypassed, the user has better flatness, but will be subject to spurs from out of band interfering signals. When the Low Noise Path is enabled, the analyzer automatically switches around certain circuitry in the high frequency bands which can contribute to noise, when it is appropriate based on other analyzer settings.

For most applications, the preset state is Standard Path, which gives the best remote-control throughput, minimizes acoustic noise from switching and minimizes the risk of wearout in the hardware switches.

For applications that utilize the wideband IF paths, the preset state is the µW Preselector Bypass path, if option MPB is present. This is because, when using a wideband IF such as the 140 MHz IF, the µW

Preselector’s bandwidth can be narrower than the available IF bandwidth, causing degraded amplitude flatness and phase linearity, so it is desirable to bypass the preselector in the default case.

Users may choose Low Noise Path Enable. It gives a lower noise floor, especially in the 21–26.5 GHz region, though without improving many measures of dynamic range, and without giving the best possible noise floor. The preamp, if purchased and used, gives better noise floor than does the Low

Noise Path, however its compression threshold and third-order intercept are much poorer than that of the non-preamp Low Noise Path. There are some applications, typically for signals around 30 dBm, for which the third-order dynamic range of the standard path is good enough, but the noise floor is not low enough even with 0 dB input attenuation. When the third-order dynamic range of the preamp path is too little and the noise floor of the standard path is too high, the Low Noise Path can provide the best dynamic range.

Key Path

Mode

Scope

Amplitude

SA, BASIC, PNOISE, VSA

Meas Global

27

Common Measurement Functions 1

Remote Command

Example

Notes

Dependencies

Preset

State Saved

Readback

Initial S/W Revision

[:SENSe]:POWer[:RF]:MW:PATH STD|LNPath|MPBypass

[:SENSe]:POWer[:RF]:MW:PATH?

:POW:MW:PATH LNP ! Enables the Low Noise path

If a Presel Center is performed, the analyzer will momentarily switch to the

Standard Path, regardless of the setting of

µW Path Control

The DC Block will always be switched in when the low noise path is switched in, to protect succeeding circuitry from DC. Note that this does not mean

“when the low noise path is enabled” but when, based on the Low Noise Path rules, the path is actually switched in. This can happen when the selection is

Low Noise Path Enable

. In the case where the DC Block is switched in the analyzer is now AC coupled. However, if the user has selected DC coupling, the UI will still behave as though it were DC coupled, including all annunciation, warnings, status bits, and responses to SCPI queries. This is because, based on other settings, the analyzer could switch out the low noise path at any time and hence go back to being DC coupled.

Alignment switching ignores the settings in this menu, and restores them when finished.

Blanked in BBIQ

All modes other than IQ Analyzer and VXA modes: STD

IQ Analyzer and VXA mode:

MPB option present and licensed: MPB

MPB option not present and licensed: STD

Save in State

Value selected in the submenu

A.04.00

Standard Path

This path gives the best remote-control throughput, minimizes acoustic noise from switching and minimizes the risk of wear in the hardware switches, particularly in remote test scenarios where both low band and high band setups will follow in rapid succession.

In this path, the bypass of the low band/high band switch and microwave preamp is never activated, which can cause some noise degradation but preserves the life of the bypass switch.

Key Path

Example

Readback Text

Initial S/W Revision

Amplitude, µW Path Control

:POW:MW:PATH STD

Standard Path

A.04.00

Low Noise Path Enable

You may choose Low Noise Path Enable, which gives a lower noise floor under some circumstances,

28

Common Measurement Functions 1

particularly when operating in the 21–26.5 GHz region. With the Low Noise Path enabled, the low band/high band switch and microwave preamp are bypassed whenever all of the following are true:

• The analyzer is not in the Low Band, meaning:

• the start frequency is above 3.5 GHz and

• the stop frequency is above 3.6 GHz.

• the internal preamp is not installed or (if installed) is set to

Off

or

Low Band

Note that this means that, when any part of a sweep is done in Low Band, the Low Noise Path is not used, whether or not the

Low Noise Path Enable

is selected in the user interface. Also, if the preamp is turned on, the Low Noise Path is not used, whether or not the

Low Noise Path Enable

is selected in the user interface. The only time the Low Noise Path is used is when

Low Noise Path Enable

is selected, the sweep is completely in High Band (> 3.6 GHz) and no preamp is in use.

See “More Information” on page 29

Key Path

Measurement

Example

Notes

Dependencies

Readback Text

Initial S/W Revision

Amplitude, µW Path Control

Swept SA

:POW:MW:PATH LNP

For measurements that use IQ acquisition, the low noise path is used when the

Center Frequency is in High Band (> 3.6 GHz) and no preamp is in use.

In other words, the rules above are modified to use only the center frequency to qualify which path to switch in.

This is not the case for FFT’s in the Swept SA measurement; they use the same rules as swept measurements.

Key is blanked if current mode does not support it.

Key is grayed out if mode supports it but current measurement does not support it.

Unless Option LNP is present and licensed, key is blank and if SCPI command sent, error –241, "Hardware missing; Option not installed" is generated.

Low Noise Path Enable

A.04.00

More Information

The user should understand that the Low Noise Path, while giving improved DANL, has the disadvantage of decreased TOI performance and decreased gain compression performance relative to the standard path.

The user should also understand that the bypass switch is a mechanical switch and has finite life; so if the

Low Noise Path

is enabled, it is possible to cause frequent cycling of this switch by frequently changing analyzer settings such that the above conditions hold true only some of the time. A user making tests of this nature should consider opting for the

Standard Path

, which will never throw the bypass switch, at the expense of some degraded noise performance.

29

Common Measurement Functions 1

The low noise path is useful for situations where the signal level is so low that the analyzer performance is dominated by noise even with 0 dB attenuation, but still high enough that the preamp option would have excessive third-order intermodulation or compression. The preamp, if purchased and used, gives better noise floor than does the “Low Noise Path.” However, its compression threshold and third-order intercept are much poorer than that of the non-preamp path. There are some applications, typically for signals around 30 dBm, for which the third-order dynamic range of the standard path is good enough, but the noise floor is not low enough even with 0 dB input attenuation. When the third-order dynamic range of the preamp path is too little and the noise floor of the standard path is too high, the Low Noise Path can provide the best dynamic range

The graph below illustrates the concept. It shows, in red, the performance of an analyzer at different attenuation settings, both with the preamp on and off, in a measurement that is affected by both analyzer noise and analyzer TOI. The green shows the best available dynamic range, offset by 0.5 dB for clarity.

The blue shows how the best available dynamic range improves for moderate signal levels with the low noise path switched in. In this illustration, the preamp improves the noise floor by 15 dB while degrading the third-order intercept by 30 dB, and the low noise path reduces loss by 8 dB. The attenuator step size is 2 dB.

There are other times where selecting the low noise path improves performance, too.

Compression-limited measurements such as finding the nulls in a pulsed-RF spectrum can profit from the low noise path in a way similar to the TOI-limited measurement illustrated. Accuracy can be improved when the low noise path allows the optimum attenuation to increase from a small amount like

0, 2 or 4 dB to a larger amount, giving better return loss at the analyzer input. Harmonic measurements, such as second and third harmonic levels, are much improved using the low noise path because of the superiority of that path for harmonic (though not intermodulation) distortion performance.

µW Preselector Bypass

This key toggles the preselector bypass switch for band 1 and higher. When the microwave presel is on, the signal path is preselected. When the microwave preselector is off, the signal path is not preselected.

30

Common Measurement Functions 1

The preselected path is the normal path for the analyzer.

The preselector is a tunable bandpass filter which prevents signals away from the frequency of interest from combining in the mixer to generate in-band spurious signals (images). The consequences of using a preselector filter are its limited bandwidth, the amplitude and phase ripple in its passband, and any amplitude and phase instability due to center frequency drift.

Option MPB or pre-selector bypass provides an unpreselected input mixer path for certain X-Series signal analyzers with frequency ranges above 3.6 GHz. This signal path allows a wider bandwidth and less amplitude variability, which is an advantage when doing modulation analysis and broadband signal analysis. The disadvantage is that, without the preselector, image signals will be displayed. Another disadvantage of bypassing the preselector is increased LO emission levels at the front panel input port.

Key Path

Example

Dependencies

Readback Text

Initial S/W Revision

Amplitude, µW Path Control

:POW:MW:PATH MPB

Key is blanked if current mode does not support it.

Key is grayed out if mode supports it but current measurement does not support it.

Key is blank unless Option MPB is present and licensed. If SCPI command sent when MPB not present, error –241, "Hardware missing; Option not installed" is generated.

µW Preselector Bypass

A.04.00

Example

Preset

Backwards Compatibility

SCPI

Backwards Compatibility

Notes

:POW:MW:PRES OFF

Bypasses the microwave preselector

ON

[:SENSe]:POWer[:RF]:MW:PRESelector[:STATe] ON|OFF|0|1

[:SENSe]:POWer[:RF]:MW:PRESelector[:STATe]?

The ON parameter sets the STD path. The OFF parameter sets Path MPB.

Internal Preamp

Accesses a menu of keys that control the internal preamps. Turning on the preamp gives a better noise figure, but a poorer TOI to noise floor dynamic range. You can optimize this setting for your particular measurement.

The instrument takes the preamp gain into account as it sweeps. If you sweep outside of the range of the preamp the instrument will also account for that. The displayed result will always reflect the correct gain.

Key Path

Scope

AMPTD Y Scale

Meas Global

31

Common Measurement Functions 1

Remote Command

Dependencies

Preset

State Saved

Initial S/W Revision

[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1

[:SENSe]:POWer[:RF]:GAIN[:STATe]?

Preamp is not available on all hardware platforms. If the preamp is not present or is unlicensed, the key is not shown.

The preamp is not available when the electronic/soft attenuator is enabled.

OFF

Saved in instrument state

Prior to A.02.00

Key Path

Scope

Remote Command

Dependencies

AMPTD Y Scale, Internal Preamp

Meas Global

[:SENSe]:POWer[:RF]:GAIN:BAND LOW|FULL

[:SENSe]:POWer[:RF]:GAIN:BAND?

Preamp is not available on all hardware platforms. If the preamp is not present or is unlicensed, the key is not shown.

If a POW:GAIN:BAND FULL command is sent when a low band preamp is available, the preamp band parameter is to LOW instead of FULL, and an

"Option not installed" message is generated.

LOW

Saved in instrument state

Prior to A.02.00

Preset

State Saved

Initial S/W Revision

Off

Turns the internal preamp off

Key Path

Example

Readback

Initial S/W Revision

AMPTD Y Scale, Internal Preamp

:POW:GAIN OFF

Off

Prior to A.02.00

Low Band

Sets the internal preamp to use only the low band.

The frequency range of the installed (optional) low-band preamp is displayed in square brackets on the

Low Band

key label.

Key Path

AMPTD Y Scale, Internal Preamp

32

Common Measurement Functions 1

Example

Readback

Initial S/W Revision

:POW:GAIN ON

:POW:GAIN:BAND LOW

Low Band

Prior to A.02.00

Full Range

Sets the internal preamp to use its full range. The low band (0–3.6 GHz or 0–3GHz, depending on the model) is supplied by the low band preamp and the frequencies above low band are supplied by the high band preamp.

The frequency range of the installed (optional) preamp is displayed in square brackets on the

Full Range

key label. If the high band option is not installed the Full Range key does not appear.

Key Path

Example

Readback

Initial S/W Revision

AMPTD Y Scale, Internal Preamp

:POW:GAIN ON

:POW:GAIN:BAND FULL

Full Range

Prior to A.02.00

Auto Couple

The Auto Couple feature provides a quick and convenient way to automatically couple multiple instrument settings. This helps ensure accurate measurements and optimum dynamic range. When the

Auto Couple feature is activated, either from the front panel or remotely, all parameters of the current measurement which have an Auto/Manual mode are set to Auto mode and all measurement settings dependent on (or coupled to) the Auto/Man parameters are automatically adjusted for optimal performance.

However, the Auto Couple keyactions are confined to the current measurement only. It does not affect other measurements in the mode, and it does not affect markers, marker functions, or trace or display

More Informationattributes.

See “More Information” on page 34

Key Path

Remote Command

Example

Notes

Initial S/W Revision

Front-panel key

:COUPle ALL|NONE

:COUP ALL

:COUPle ALL puts all Auto/Man parameters in Auto mode (equivalent to pressing the

Auto Couple

key).

:COUPLE NONE puts all Auto/Man parameters in manual mode. It decouples all the coupled instrument parameters and is not recommended for making measurements.

Prior to A.02.00

33

Common Measurement Functions 1

More Information

There are two types of functions that have Auto/Manual modes.

Auto/Man Active Function keys

An Auto/Man toggle key controls the binary state associated with an instrument parameter by toggling between "Auto" (where the parameter is automatically coupled to the other parameters it is dependent upon) and "Man" (where the parameter is controlled independent of the other parameters), as well as making the parameter the active function. The current mode is indicated on the softkey with either

"Auto" or "Man" underlined as illustrated below.

Sweep Time

66.24 ms

Auto Man vsd07

Auto/Man 1-of-N keys

An Auto/Man 1-of-N key allows you to manually pick from a list of parameter values, or place the function in "Auto" in which case the value is automatically selected (and indicated) as shown below. If in

Auto, Auto is underlined on the calling key. If in manual operation, manual is indicated on the calling key. But the calling key does not actually toggle the function, it simply opens the menu.

X Axis Scale

Frequency

Auto Man

X Axis Scale

Auto

Frequency

Period

Time

Inverse Time

vsd08

34

Common Measurement Functions 1

Cont (Continuous Measurement/Sweep)

Sets the analyzer for Continuous measurement operation. The single/continuous state is Meas Global so the setting will affect all measurements. If you are Paused, pressing

Cont

does a Resume.

Remote Command:

Example:

:INITiate:CONTinuous OFF|ON|0|1

:INITiate:CONTinuous?

:INIT:CONT 0 puts analyzer in Single measurement operation.

:INIT:CONT 1 puts analyzer in Continuous measurement operation

Preset:

State Saved:

Key Path:

Initial S/W Revision:

ON

(Note that SYST:PRESet sets INIT:CONT to ON but *RST sets INIT:CONT to OFF)

Saved in Instrument State

Front-panel key

Prior to A.02.00

In Swept SA Measurement (Spectrum Analysis Mode):

The analyzer takes repetitive sweeps, averages, measurements, etc, when in Continuous mode. When the average count reaches the

Average/Hold Number

the count stops incrementing, but the analyzer keeps sweeping. See the Trace/Detector section for the averaging formula used both before and after the

Average/Hold Number

is reached. The trigger condition must be met prior to each sweep. The type of trace processing for multiple sweeps, is set under the Trace/Detector key, with choices of

Trace

Average

,

Max Hold

, or

Min Hold

.

In Other Measurements/Modes:

With

Avg Number

(in the

Meas Setup

menu) set to

Off

or set to

On

with a value of 1, a sweep is taken after the trigger condition is met; and the analyzer continues to take new sweeps after the current sweep has completed and the trigger condition is again met. However, with

Avg Number

set to On with a value

>1, multiple sweeps (data acquisitions) are taken for the measurement. The trigger condition must be met prior to each sweep. The sweep is not stopped when the average count k equals the number N set for Avg

Number is reached, but the number k stops incrementing. A measurement average usually applies to all traces, marker results, and numeric results. But sometimes it only applies to the numeric results.

If the analyzer is in Single measurement, pressing the

Continuous

key does not change k and does not cause the sweep to be reset; the only action is to put the analyzer into Continuous measurement operation.

If it is already in continuous sweep:

• the INIT:CONT 1 command has no effect

• the INIT:CONT 0 command will place the analyzer in Single Sweep but will have no effect on the current sequence until k = N, at which point the current sequence will stop and the instrument will go

35

Common Measurement Functions 1

to the idle state.

FREQChannel

Accesses a menu of keys that allow you to control the Frequency and Channel parameters of the instrument.

Some features in the Frequency menu are the same for all measurements – they do not change as you change measurements. Settings like these are called “Meas Global” and are unaffected by Meas Preset.

For example, the Center Freq setting is the same for all measurements - it does not change as you change measurements.

Key Path

Initial S/W Revision

Front-panel key

Prior to A.02.00

Auto Tune

Auto Tune is an immediate action key. When it is pressed, it causes the analyzer to change Center

Frequency to the strongest signal in the tunable span of the analyzer, excluding the LO. It is designed to quickly get you to the most likely signal(s) of interest, with no signal analysis knowledge required. As such, there are no configurable parameters for this feature. There are only pre-selected values that work in most real world situations.

NOTE

You may see a slight pause before the signal of interest is presented at midscreen.

Remote Command:

Initial S/W Revision:

Key Path:

[:SENSe]:FREQuency:TUNE:IMMediate

Prior to A.02.00

FREQ Channel

Center Freq

Sets the frequency that corresponds to the horizontal center of the graticule (when frequency Scale Type is set to linear). While adjusting the Center Frequency the Span is held constant, which means that both

Start Frequency and Stop Frequency will change.

Pressing Center Freq also sets the frequency entry mode to Center/Span. In Center/Span mode, the center frequency and span values are displayed below the graticule, and the default active function in the

Frequency menu is

Center Freq

.

The center frequency setting is the same for all measurements within a mode, that is, it is Meas Global.

Some modes are also able to share a Mode Global center frequency value. If this is the case, the Mode will have a

Global Settings

key in its

Mode Setup

menu.

The

Center Freq

function sets (and queries) the Center Frequency for the currently selected input. If your analyzer has multiple inputs, and you select another input, the Center Freq changes to the value for that input. SCPI commands are available to directly set the Center Freq for a specific input.

36

Common Measurement Functions 1

See

“RF Center Freq” on page 39

See

“I/Q Center Freq” on page 39

See

“Center Frequency Presets” on page 38

Key Path

Scope

Remote Command

Example

Notes

Dependencies

Couplings

Preset

State Saved

Min

Max

FREQ Channel

Meas Global

[:SENSe]:FREQuency:CENTer <freq>

[:SENSe]:FREQuency:CENTer?

FREQ:CENT 50 MHz

FREQ:CENT UP !changes the center frequency to 150 MHz if you use

FREQ:CENT:STEP 100 MHz !to set the center frequency step size to 100

MHz

FREQ:CENT?

This command sets either the RF or I/Q Center Frequency depending on the selected input.

For RF input it is equivalent to FREQ:RF:CENT

For I/Q input it is equivalent to FREQ:IQ:CENT

Preset and Max values are dependant on Hardware Options (503, 507, 508,

513, 526)

The Center Frequency can be limited by Start or Stop Freq limits, if the Span is so large that Start or Stop hit their limit.

When operating in “swept span”, any value of the Center Frequency or Span that is within the frequency range of the analyzer is allowed when the value is being set through the front panel numeric key pad or the SCPI command. The other parameter is forced to a different value if needed, to keep the Start and the Stop Frequencies within the analyzer’s frequency range

Depends on instrument maximum frequency, mode, measurement, and selected input..

See

“Center Frequency Presets” on page 38 ,

“RF Center Freq” on page 39 ,

and

“I/Q Center Freq” on page 39 .

Saved in instrument state

Depends on instrument maximum frequency, mode, measurement, and selected input..

See

“Center Frequency Presets” on page 38 ,

“RF Center Freq” on page 39 ,

and

“I/Q Center Freq” on page 39 .

Depends on instrument maximum frequency, mode, measurement, and selected input..

See

“Center Frequency Presets” on page 38 ,

“RF Center Freq” on page 39 ,

and

“I/Q Center Freq” on page 39 .

37

Common Measurement Functions 1

Freq Option

503

507

508

513

526

Default Unit

Status Bits/OPC Dependencies

Initial S/W Revision

Hz non-overlapped

Prior to A.02.00

Center Frequency Presets

Model numbers N9010A, N9020A, N9030A:

CF after Mode

Preset

1.805 GHz

3.505 GHz

4.205 GHz

6.805 GHz

13.255 GHz

Stop Freq after

Mode Preset

3.6 GHz

7.0 GHz

8.4 GHz

13.6 GHz

26.5 GHz

Max Freq

(can't tune above)

3.7 GHz

7.1 GHz

8.5 GHz

13.8 GHz

27.0 GHz

Model number N9000A.

Freq Option

503

507

CF after Mode

Preset

1.505 GHz

3.755 GHz

Stop Freq after

Mode Preset

3.0 GHz

7.5 GHz

Max Freq

(can't tune above)

3.08 GHz

7.58 GHz

The following table shows the Center Frequency Presets for modes other than Spectrum Analyzer:

Mode

WCDMA

WIMAXOFDMA,

BASIC

ADEMOD

VSA

TDSCDMA

PNOISE

GSM

NFIGURE

CF Preset for RF

1 GHz

1 GHz

1 GHz

1 GHz

1 GHz

1 GHz

1 GHz

935.2 MHz

1.505 GHz

38

Common Measurement Functions 1

RF Center Freq

SCPI command for specifying the RF Center Frequency. This command will set the Center Frequency to be used when the RF input is selected, even if the RF input is not the input which is selected at the time the command is sent. Note that the

Center Freq

function in the

Frequency

menu on the front panel always applies to the currently selected input.

Scope

Remote Command

Example

Notes

Dependencies

Preset

State Saved

Min

Max

Meas Global

[:SENSe]:FREQuency:RF:CENTer <freq>

[:SENSe]:FREQuency:RF:CENTer?

FREQ:RF:CENT 30 MHz

This command is the same in all modes, but the parameter is

Measurement Global. So the value is independent in each mode and common across all the measurements in the mode.

If the electronic/soft attenuator is enabled, any attempt to set Center

Frequency such that the Stop Frequency would be >3.6 GHz fails and results in an advisory message. If the equivalent SCPI command is sent, this same message is generated as part of a “–221, Settings conflict” warning.

See table above

Saved in instrument state.

–79.999995 MHz

See table above. Basically instrument maximum frequency – 10 Hz minimum span. If the knob or step keys are being used, depends on the value of the other three interdependent parameters

Prior to A.02.00

A.03.00

Initial S/W Revision

Modified at S/W

Revision

I/Q Center Freq

SCPI command for specifying the I/Q Center Frequency. This command will set the Center Frequency to be used when the I/Q input is selected, even if the I/Q input is not the input which is selected at the time the command is sent. Note that the

Center Freq

function in the

Frequency

menu on the front panel always applies to the currently selected input.

Scope

Remote Command

Example

Meas Global

[:SENSe]:FREQuency:IQ:CENTer <freq>

[:SENSe]:FREQuency:IQ:CENTer?

FREQ:IQ:CENT: 30 MHz

39

Common Measurement Functions 1

Notes

Preset

State Saved

Min

Max

Initial S/W Revision

This command is the same in all modes, but the parameter is

Measurement Global. So the value is independent in each mode and common across all the measurements in the mode.

0 Hz

Saved in instrument state.

–40.049995 MHz

40.049995 MHz

Prior to A.02.00

CF Step

Changes the step size for the center frequency and start and stop frequency functions. Once a step size has been selected and the center frequency function is active, the step keys (and the UP|DOWN parameters for Center Frequency from remote commands) change the center frequency by the step-size value. The step size function is useful for finding harmonics and sidebands beyond the current frequency span of the analyzer.

Note that the start and stop frequencies also step by the CF Step value.

Key Path

Remote Command

Example

Notes

Dependencies

Couplings

FREQ Channel

[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>

[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?

[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1

[:SENSe]:FREQuency:CENTer:STEP:AUTO?

FREQ:CENT:STEP:AUTO ON

FREQ:CENT:STEP 500 MHz

FREQ:CENT UP !increases the current center frequency value by 500 MHz

FREQ:CENT:STEP?

FREQ:CENT:STEP:AUTO?

Preset and Max values are dependant on Hardware Options (503, 508, 513,

526)

Span, RBW, Center frequency

If the electronic/soft attenuator is enabled, any attempt to change the value of the center frequency >3.6 GHz by pressing the Up-arrow key, fails and results in an advisory message. If the equivalent SCPI command is sent, this same message is generated as part of a “–221, Settings conflict” warning.

When auto-coupled in a non-zero span, the center frequency step size is set to

10% of the span. When auto-coupled in zero span, the center frequency step size is set to the equivalent –3 dB RBW value.

40

Common Measurement Functions 1

Preset

State Saved

Min

Max

Default Unit

Status Bits/OPC dependencies

Initial S/W Revision

Modified at S/W Revision

Auto

ADEMOD: 1 MHz

ON

Saved in instrument state

– ( the maximum frequency of the instrument). (that is, 27 GHz max freq instrument has a CF step range of +/– 27 GHz) the maximum frequency of the instrument. (that is, 27 GHz max freq instrument has a CF step range of +/– 27 GHz)

Hz non-overlapped

Prior to A.02.00

A.03.00

Freq Offset

Enables you to set a frequency offset value to account for frequency conversions outside of the analyzer.

This value is added to the display readout of the marker frequency, center frequency, start frequency, stop frequency, and all other absolute frequency settings in the analyzer including frequency count. When a frequency offset is entered, the value appears below the center of the graticule. To eliminate an offset, perform a Mode Preset or set the frequency offset to 0 Hz.

See “More Information” on page 42

.

Remote Command

Example

Default Unit

Initial S/W Revision

Key Path

Scope

Notes

Dependencies

Preset

State Saved

Min

Max

Status Bits/OPC dependencies

[:SENSe]:FREQuency:OFFSet <freq>

[:SENSe]:FREQuency:OFFSet?

FREQ:OFFS 10 MHz

Hz

Prior to A.02.00

FREQ Channel

Meas Global

Preset and Max values are depending on Hardware Options (503, 507, 508,

513, 526)

Frequency Offset is not available when the frequency scale is set to Log

See the table in See

“Center Frequency Presets” on page 38

Saved in instrument state

–500 GHz

500 GHz non-overlapped

41

Common Measurement Functions 1

Backwards Compatibility SCPI

Modified at S/W Revision

DISPlay:WINDow[1]:TRACe:X[:SCALe]:OFFSet

A.04.00

More Information

This command does not affect any bandwidths or the settings of relative frequency parameters such as delta markers or span. It does not affect the current hardware settings of the analyzer, but only the displayed frequency values. Entering an offset does not affect the trace position or display, just the value of the start and stop frequency and the values represented by the trace data. The frequency values of exported trace data, queried trace data, markers, trace data used in calculations such as N dB points, trace math, etc., are all affected by Freq Offset. Changing the offset, even on a trace that is not updating will immediately change all of the above, without taking new data.

NOTE

If a trace is exported with a nonzero Freq Offset, the exported data will contain the trace data with the offset applied. Therefore, if that trace were to be imported back into the analyzer, you would want Freq Offset to be 0, or the offset would be applied again to data which is already offset. No such care need be taken when saving a State+Trace file because the data and state are saved together.

42

Common Measurement Functions 1

Input/Output

The Input/Output features are common across multiple Modes and Measurements. These common features are described in this section. See the Measurement description for information on features that are unique.

The Input/Output key accesses the softkeys that control the Input/Output parameters of the instrument.

In general, these are functions associated with external connections to the analyzer, either to the inputs or the outputs. Since these connections tend to be fairly stable within a given setup, in general the input/output settings do not change when you Preset the analyzer.

Other functions related to the input/output connections, but which tend to change on a measurement by measurement basis, can be found under

Trigger

and

AMPTD Y Scale.

In addition, some of the digital

I/O bus configurations can be found under

System

.

NOTE

The functions in the Input/Output menu are "global" (common) to all Modes

(applications). But individual Input/Output functions only appear in a Mode if they apply to that Mode. Functions that apply to a Mode but not to all measurements in the Mode may be grayed-out in some measurements.

See “Input/Output variables - Preset behavior” on page 43

The Input Port selection is the first menu under

Input/Output

:

Remote Command:

[:SENSe]:FEED

RF|AIQ|IQ|IONLy|QONLy|INDependent|AREFerence

[:SENSe]:FEED?

Backwards Compatibility SCPI: The legacy parameter [:SENSe]:FEED AREFerence is aliased to the new command [:SENSe]:FEED:AREF REF50 for backwards compatibility. This causes the input to be switched to the 50 MHz calibrator.

Preset:

State Saved:

Key Path:

Initial S/W Revision:

This setting is unaffected by a Preset or power cycle. It survives a Mode Preset and mode changes.

It is set to RF on a "Restore Input/Output Defaults" or "Restore System

Defaults->All"

Saved in instrument state

Front-panel key

Prior to A.02.00

Input/Output variables - Preset behavior

Virtually all the input/output settings are NOT a part of mode preset. They can be set to their default value by one of the three ways - by using the Restore Input/Output Defaults key on the first page of the input/output menu, by using the System->Restore System Defaults->Input/Output Settings or by using the System -> Restore System Defaults->All. Also, they survive Preset and Power cycle.

43

Common Measurement Functions 1

A very few of the Input/Output settings do respond to a Mode Preset; for example, if the Calibrator is on it turns off on a Preset, and if DC coupling is in effect it switches to AC on a Preset. These exceptions are made in the interest of reliability and usability, which overrides the need for absolute consistency.

Exceptions are noted in the SCPI table for the excepted functions.

RF Input

Selects the front-panel RF input port to be the analyzer signal input. If RF is already selected, pressing this key accesses the RF input setup functions.

Example:

Key Path:

Readback:

Initial S/W Revision:

[:SENSe]:FEED RF

Input/Output

The current input impedance settings are Readback to this key i.e.

"XX, ZZ" where XX is AC or DC and ZZ is 50 or 75

Prior to A.02.00

Input Z Correction

Sets the input impedance for unit conversions. This affects the results when the y-axis unit is voltage or current units (dBmV, dBµV, dBµA, V, A), but not when it is power units (dBm, W). The impedance you select is for computational purposes only, since the actual impedance is set by internal hardware to 50 ohms. Setting the computational input impedance to 75 ohms is useful when using a 75 ohm to 50 ohm adapter to measure a 75 ohm device on an analyzer with a 50 ohm input impedance.

There are a variety ways to make 50 to 75 ohm transitions, such as impedance transformers or minimum loss pads. The choice of the solution that is best for your measurement situation requires balancing the amount of loss that you can tolerate with the amount of measurement frequency range that you need. If you are using one of these pads/adaptors with the

Input Z Corr

function, you might also want to use the

Ext Gain

key. This function is used to set a correction value to compensate for the gain (loss) through your pad. This correction factor is applied to the displayed measurement values.

Remote Command:

[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude] 50|75

[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]?

Example:

Preset:

CORR:IMP 75 sets the input impedance correction to 75 ohms.

CORR:IMP?

This is unaffected by a Preset but is set to 50 ohms on a "Restore Input/Output

Defaults" or "Restore System Defaults->All"

Some instruments/options may have 75 ohms available.

State Saved:

Key Path:

Readback:

44

Saved in instrument state

Input/Output, RF Input

50

Ω or 75 Ω. Current setting reads back to the RF key.

Common Measurement Functions 1

Initial S/W Revision: Prior to A.02.00

RF Coupling

Specifies alternating current (AC) or direct current (DC) coupling at the analyzer RF input port.

Selecting AC coupling switches in a blocking capacitor that blocks any DC voltage present at the analyzer input. This decreases the input frequency range of the analyzer, but prevents damage to the input circuitry of the analyzer if there is a DC voltage present at the RF input.

In AC coupling mode, you can view signals below the corner frequency of the DC block, but below a certain frequency the amplitude accuracy is not specified. The frequency below which specifications do not apply is:

X-Series Model Lowest Freq for meeting specs when AC coupled

N9000A

N9010A

N9020A

N9030A

100 kHz

10 MHz

10 MHz

10 MHz

Lowest Freq fo specs when DC n/a

9 kHz

3 Hz

3 Hz

Some amplitude specifications apply only when coupling is set to DC. Refer to the appropriate amplitude specifications and characteristics for your analyzer.

When operating in DC coupled mode, ensure protection of the analyzer input circuitry by limiting the

DC part of the input level to within 200 mV of 0 Vdc. In AC or DC coupling, limit the input RF power to

+30 dBm (1 Watt).

Remote Command:

Example:

Dependencies:

Preset:

State Saved:

Key Path:

Initial S/W Revision:

Modified at S/W Revision:

:INPut:COUPling AC|DC

:INPut:COUPling?

INP:COUP DC

This key does not appear in models that are always AC coupled. When the

SCPI command to set DC coupling is sent to these models, it results in the error “Illegal parameter value;This model is always AC coupled” In these models, the SCPI query INP:COUP? always returns AC.

AC

Saved in instrument state.

Input/Output, RF Input

Prior to A.02.00

A.03.00

I/Q

This feature is not available unless the “Baseband I/Q (Option BBA)” on page 46

is installed.

45

Common Measurement Functions 1

Selects the front-panel I/Q input ports to be the analyzer signal input. If I/Q is already selected, pressing this key accesses the I/Q setup menu.

Example

Initial S/W Revision

Key Path

Mode

Notes

Notes

FEED AIQ

Prior to A.02.00

Input/Output

BASIC, CDMA2K, EDGEGSM, TDSCMDA, VSA89601,

WIMAXOFDMA

Not all measurements support the use of the I/Q signal input. When I/Q is selected in a measurement that does not support it, the "Meas invalid with

I/Q inputs" error condition occurs.

The parameters IQ|IONLy|QONLy are only supported for backwards compatibility The E44406 SCPI has the following that corresponds to

FEED:IQ:TYPE for X-Series.

[:SENSe]:FEED IQ|IONLy|QONLy

[:SENSe]:FEED?

[:SENSe]:FEED IQ will set the I/Q path to IQ

[:SENSe]:FEED IONLy will set the I/Q path to I Only

[:SENSe]:FEED QONLy will set the I/Q path to QOnly

Note [:SENSe]:FEED? will not be backward compatible.

The query [:SENSe]:FEED? will always returns AIQ whatever the type of legacy parameters IQ|IONLy|QONLy has been used.

Baseband I/Q (Option BBA)

The Baseband I/Q functionality is a hardware option. It is option BBA. If the option is not installed, none of the I/Q functionality is enabled.

The Baseband I/Q has four input ports and one output port. The input ports are I, I-bar, Q, and Q-bar. The

I and I-bar together compose the I channel and the Q and Q-bar together compose the Q channel. Each channel has two modes of operation, Single-Ended (also called "unbalanced") and Differential Input

(also called "balanced"). When in Single-Ended operation, only the main port (I or Q) is used and the complementary port (I-bar or Q-bar) is ignored. When in Differential Input mode, both main and complementary ports are used.

The input settings (range, attenuation, skew, impedance, external gain) apply to the channels, not the individual ports.

The system supports a variety of 1 M

Ω input passive probes as well as the Agilent 113x Series active differential probes using the Infinimax probe interface.

The Agilent 113x Series active probes can be used for both single ended and differential measurements.

In either case a single connection is made for each channel (on either the I or Q input). The input is automatically configured to 50

Ω single ended and the probe power is supplied through the Infinimax interface. The probe can be configured for a variety of input coupling and low frequency rejection

46

Common Measurement Functions 1

modes. In addition, a wide range of offset voltages and probe attenuation accessories are supported at the probe interface. The active probe has the advantage that it does not significantly load the circuit under test, even with unity gain probing.

With passive 1 M

Ω probes, the probe will introduce a capacitive load on the circuit, unless higher attenuation is used at the probe interface. Higher attenuation reduces the signal level and degrades the signal-to-noise-ratio of the measurement. Passive probes are available with a variety of attenuation values for a moderate cost. Most Agilent passive probes can be automatically identified by the system, setting the input impedance setting required as well as the nominal attenuation. For single ended measurements a single probe is used for each channel. Other passive probes can by used, with the attenuation and impedance settings configured manually.

For full differential measurements, the system supports probes on each of the four inputs. The attenuation of the probes should be the same for good common mode rejection and channel match.

Both active and passive probes in single ended and differential configurations can be calibrated. This calibration uses the Cal Out BNC connection and a probe connection accessory. The calibration achieves excellent absolute gain flatness in a probed measurement. It matches both the gain and frequency response of the I and Q channels as well as any delay skew, resulting in high accuracy in derived measurements such as Error Vector Magnitude (EVM).

When a probe is connected a status message will be displayed. The message will indicate if calibration data is available or not. Calibration data is saved for each type of probe (including "none") for each port and will be reapplied whenever that type of probe is re-connected to the same port. For probes with

EEPROM identification, the calibration data will be stored based on the unique probe identifier and will reapply data for that particular probe if it is available. The data will not follow a probe from one port to another. For probes without EEPROM identification, the instrument cannot distinguish between different probes of the same type and it will use the data from the last calibration for that probe type on that port.

When in differential mode, both the main and complementary probes are expected to be of the same type.

In some situations, the I and Q channels should be configured identically. In other situations it is convenient to control them independently. Some menus have a "Q Same as I" setting that will cause the

Q channel configuration to mirror the I channel configuration, avoiding the overhead of double data entry when the channels should be the same.

The output port is for calibrating the I/Q input ports, although it can also be manually controlled.

There are two types of calibrations available: cable calibration and probe calibration. The cable calibration will guide the user through connecting each input port in turn. All ports must be calibrated together. The probe calibration is done for a specific channel (I or Q). If in Single-Ended mode, only the main port is calibrated. When in Differential Input mode, the user is guided through calibrating both main and complementary ports.

The front panel I/Q port LEDs indicate the current state of that port. On (green) indicates it is active, and off (dark) indicates it is not in use. For example, the Cal Out port LED is on if and only if there is signal coming out of that port.

The input is a context and some parameters have separate values for each context. The SCPI for these parameters has an optional "[:RF|IQ]" node. If the specific context is omitted, the command acts on the current input context's value. Here are the parameters that are input context sensitive:

• Center Frequency

47

Common Measurement Functions 1

• Trigger Source

It is important to distinguish between the I and Q input ports and the displayed I and Q data values. The I and Q input ports feed into a digital receiver that does digital tuning and filtering. The I and Q data seen by the user (either on the display or through SCPI) corresponds to the real ("I") and the imaginary ("Q") output from the digital receiver. When the input path is I+jQ or I Only and the center frequency is 0 Hz the I input ends up in as the real output from the receiver and appears as "I" data. Likewise, when the input path is I+jQ and the center frequency is 0 Hz, the Q input ends up as the imaginary output from the receiver and appears as "Q" data. However, when the input path is Q Only, the Q input is sent to the receiver as Q+j0, so the receiver output has the Q input coming out on the real output, and so in Q Only, the signal from the Q input port appears as the "I" data. Another situation where the I and Q data do not necessarily correspond directly to the I and Q inputs is when the center frequency is non-zero. The digital processing involved in the tuning is a complex operation. This will result in I Only data appearing as both "I" and "Q" data, the same as that signal would appear if seen through the RF input port.

I/Q Path

Selects which I/Q input channels are active. The LED next to each I/Q input port will be on when that port is active.

The analysis bandwidth for each channel is the same as that of the instrument. So, for example, the base

N9020A has a bandwidth of 10 MHz. With I/Q input the I and Q channels would each have an analysis bandwidth of 10 MHz, giving 20 MHz of bandwidth when the I/Q Path is I+jQ. With option B25, the available bandwidth becomes 25 MHz, giving 25 MHz each to I and Q and 50 MHz to I+jQ.

I/Q voltage to power conversion processing is dependent on the I/Q Path selected.

• With I+jQ input we know that the input signal may not be symmetrical about 0 Hz, because it has a complex component. Therefore, above 0 Hz only the positive frequency information is displayed, and below 0 Hz only the negative frequency information is displayed.

• With all other Input Path selections, the input signal has no complex component and therefore is always symmetrical about 0 Hz. In this case, by convention, the power conversion shows the combined voltage for both the positive and negative frequencies. The information displayed below 0

Hz is the mirror of the information displayed above 0 Hz. This results in a power reading 6.02 dB higher (for both) than would be seen with only the positive frequency voltage. Note also that, in this case the real signal may have complex modulation embedded in it, but that must be recovered by further signal processing.

Remote Command

Example

Initial S/W Revision

Key Path

Notes

Preset

[:SENSe]:FEED:IQ:TYPE IQ|IONLy|QONLy|INDependent

[:SENSe]:FEED:IQ:TYPE?

Set the input to be both the I and Q channels, combined as I + j * Q.

FEED:IQ:TYPE IQ

Prior to A.02.00

Input/Output, I/Q

The Independent I and Q selection is only available in GPVSA

IQ

48

Common Measurement Functions 1

State Saved

Range

Readback Text

Yes

This is unaffected by a Preset but is set to the default value on a "Restore

Input/Output Defaults" or "Restore System Defaults->All"

I+jQ | I Only | Q Only | Independent I and Q

I+jQ | I Only | Q Only | Ind I/Q

Remote Command:

Preset:

Initial S/W Revision:

:INPut[1]:IQ:TYPE IQ|I|Q

:INPut[1]:IQ:TYPE?

IQ

Prior to A.02.00

I+jQ

Sets the signal input to be both the I and Q channels. The I and Q channel data will be combined as I + j * Q.

Key Path

Example

Input/Output, I/Q, I/Q Path

Set the input to be both the I and Q channels, combined as I + j * Q.

FEED:IQ:TYPE IQ

Prior to A.02.00

Initial S/W Revision

I Only

Sets the signal input to be only the I channel. The Q channel will be ignored. The data collected is still complex.

When the center frequency is 0 the imaginary part will always be zero, but for any other center frequency both the real and imaginary parts will be significant.

Key Path

Example

Input/Output, I/Q, I/Q Path

Set the input to be only the I channel.

FEED:IQ:TYPE IONL

Prior to A.02.00

Initial S/W Revision

Q Only

Sets the signal input to be only the Q channel. The I channel will be ignored. The Q channel will be sent to the digital receiver block as Q+j0. The receiver's output is still complex. When the center frequency is 0 the imaginary part will always be zero, but for any other center frequency both the real and imaginary parts will be significant.

Note that since the receiver's real output is displayed as the "I" data, when the center frequency is 0, the Q Only input appears as the "I" data.

Key Path

Example

Input/Output, I/Q, I/Q Path

Set the input to be only the Q channel.

FEED:IQ:TYPE QONL

49

Common Measurement Functions 1

Initial S/W Revision Prior to A.02.00

Independent I and Q

Sets the signal input to be both the I and Q channels, but as independent inputs. It is equivalent to treating I as channel 1 and Q as channel 2 in an oscilloscope. Each channel's data is still complex. When the center frequency is

0 the imaginary part will always be zero, but for any other center frequency both the real and imaginary parts will be significant.

This selection is only available in VXA.

Example

Initial S/W Revision

Key Path

Notes

Readback Text

Turn on both I and Q channels and treat I as channel 1 and Q as channel 2.

FEED:IQ:TYPE IND

Prior to A.02.00

Input/Output, I/Q, I/Q Path

The Independent I and Q selection is only available in GPVSA

Ind I/Q

I Setup

Access the channel setup parameters for the I channel.

Key Path

Initial S/W Revision

Input/Output, I/Q

Prior to A.02.00

I Differential Input

Selects differential input on or off for the I channel. For differential input (also called balanced input), the analyzer uses both main and complementary ports. When differential input is off (also called single-ended or unbalanced input), the analyzer uses only the main port.

Remote Command

Example

Initial S/W Revision

Key Path

Notes

:INPut:IQ[:I]:DIFFerential OFF|ON|0|1

:INPut:IQ[:I]:DIFFerential?

Put the I channel in Differential Input mode

INP:IQ:DIFF ON

Prior to A.02.00

Input/Output, I/Q, I Setup

When Differential Input = On, the analyzer will check for attenuation mismatches between the I and I-bar ports. If the difference in attenuation values exceeds 0.5 dB an error condition will be set.

50

Common Measurement Functions 1

Couplings

Preset

State Saved

Range

Some active probes include built-in differential capability. When one of these probes is sensed, this key is disabled. Since the differential capability is handled in the probe, the Analyzer will use only the main port and the key will show that the Analyzer's Differential Input mode is Off (indicating that the complementary port not in use).

When Q Same as I is On, the value set for I will also be copied to Q.

Off

Yes

This is unaffected by a Preset but is set to the default value on a "Restore

Input/Output Defaults" or "Restore System Defaults->All"

Off | On

Remote Command:

Preset:

Notes:

:INPut[1]:IQ:BALanced[:STATe] OFF|ON|0|1

:INPut[1]:IQ:BALanced[:STATe]?

OFF

This backwards compatibility SCPI command was for an instrument without independent settings for the I and Q channels. Therefore, it is tied only to the I channel and does not provide an equivalent for the Q channel. For proper operation of the backwards compatibility command Q Same as I should be set to On.

Prior to A.02.00

Initial S/W Revision:

I Input Z

Selects the input impedance for the I channel. The impedance applies to both the I and I-bar ports.

The input impedance controls the hardware signal path impedance match. It is not used for converting voltage to power. The voltage to power conversion always uses the Reference Z parameter. The Reference Z parameter applies to both I and Q channels.

Remote Command

Example

Initial S/W Revision

Key Path

Notes

Couplings

:INPut[1]:IQ[:I]:IMPedance LOW|HIGH

:INPut[1]:IQ[:I]:IMPedance?

Set the I channel input impedance to 1 M

Ω

INP:IQ:IMP HIGH

Prior to A.02.00

Input/Output, I/Q, I Setup

LOW = 50

Ω, HIGH = 1 MΩ

Input impedance is a built-in characteristic of a probe. Therefore, whenever a probe is sensed, this key is disabled and the value is set to match the probe.

When no probe is sensed on Q and Q Same as I is On, the value set for I will also be copied to Q.

51

Common Measurement Functions 1

Preset

State Saved

Range

LOW

Yes

This is unaffected by a Preset but is set to the default value on a "Restore

Input/Output Defaults" or "Restore System Defaults->All"

50

Ω | 1 MΩ

I Skew

Sets the skew factor for the I channel. The skew will shift the channel's data in time. Use this to compensate for differences in the electrical lengths of the input paths due to cabling.

Key Path

Remote Command

Preset

State Saved

Range

Example

Input/Output, I/Q, I Setup

[:SENSe]:CORRection:IQ[:I]:SKEW <seconds>

[:SENSe]:CORRection:IQ[:I]:SKEW?

0

Yes

This is unaffected by Preset but is set to the default value on a "Restore

Input/Output Defaults" or "Restore System Defaults->All"

0 s to 100 ns

Delay the data for the I channel by 10 ns.

CORR:IQ:SKEW 10 ns

Prior to A.02.00

Initial S/W Revision

I Probe

Access the probe setup parameters for the I channel. See

“I/Q Probe Setup” on page 56

.

Key Path

State Saved

Readback Text

Input/Output, I/Q, I Setup

No

[<I port probe id>]

This is reporting the type of probe sensed on the I port. There is no parameter for overriding what is sensed.

Prior to A.02.00

Initial S/W Revision

Combined Differential/Input Z (Remote Command Only)

This is Remote Command only (no front panel) and is for backwards compatibility only. It combines the

Differential Input and Input Z selections into a single SCPI command.

Remote Command:

:INPut:IMPedance:IQ U50|B50|U1M|B1M

:INPut:IMPedance:IQ?

52

Common Measurement Functions 1

Preset:

Notes:

Couplings:

Example:

U50

The enum values translate as follows:

U50: Differential Input = Off, Input Z = 50

Ω

B50: Differential Input = On, Input Z = 50

Ω

U1M: Differential Input = Off, Input Z = 1 M

Ω

B1M: Differential Input = On, Input Z = 1 M

Ω

This command is for backwards compatibility. It combines the Input Z (50

Ω or 1 M

Ω) parameter with the Differential Input (Off = "Unbalanced", On =

"Balanced") parameter into a single enumeration.

This backwards compatibility SCPI command was for an instrument without independent settings for the I and Q channels. Therefore, it is tied only to the I channel and does not provide an equivalent for the Q channel. For proper operation of the backwards compatibility command Q Same as I should be set to On.

Also, note the subtle difference between this SCPI command and the backwards compatibility command for Input Z. The Input Z SCPI has "IQ" before "IMP" while this command has that order reversed.

This command does not have an independent parameter, but instead is tied to the Differential Input and Input Z parameters. The coupling for those parameters apply to this command too.

:INPut:IMPedance:IQ U50

This is equivalent to the following two SCPI commands:

:INP:IQ:DIFF OFF

:INP:IQ:IMP 50

Prior to A.02.00

Initial S/W Revision:

Q Setup

Access the channel setup parameters for the Q channel.

Key Path

Readback Text

Initial S/W Revision

Input/Output, I/Q

When Q Same as I is On the readback is "Q Same as I".

Prior to A.02.00

Q Same as I

Many, but not all, usages require the I and Q channels have an identical setup. To simplify channel setup, the Q

Same as I will cause the Q channel parameters to be mirrored from the I channel. That way your only need to set up one channel (the I channel). The I channel values are copied to the Q channel, so at the time Q Same as I is turned off the I and Q channel setups will be identical. This does not apply to Probe settings or to parameters that

53

Common Measurement Functions 1

determined by the probe.

Remote Command

Example

Initial S/W Revision

Key Path

Couplings

Preset

State Saved

Range

Readback Text

:INPut:IQ:MIRRored OFF|ON|0|1

:INPut:IQ:MIRRored?

Turn off the mirroring of parameters from I to Q.

INP:IQ:MIRR OFF

Prior to A.02.00

Input/Output, I/Q, Q Setup

Only displayed for the Q channel. When Yes, the I channel values for some parameters are mirrored (copied) to the Q channel. However, when a parameter is determined by the type of probe and a probe is sensed, the probe setting is always used and the I channel setting is ignored. The following parameters are mirrored:

Differential Input (when not determined by probe)

Input Z (when not determined by probe)

This is unaffected by a Preset but is set to the default value (Q Same as I set to

"On") on a "Restore Input/Output Defaults" or "Restore System

Defaults->All"

Saved in instrument state.

On | Off

"Q Same as I" when On, otherwise none.

Q Differential Input

Selects differential input on or off for the Q channel. For differential input (also called balanced input), the analyzer uses both the Q and Q-bar ports. When differential input is off (also called single-ended or unbalanced input), the analyzer uses only the Q port.

Remote Command

Example

Initial S/W Revision

Key Path

Notes

:INPut:IQ:Q:DIFFerential OFF|ON|0|1

:INPut:IQ:Q:DIFFerential?

Put the Q channel in Differential Input mode

INP:IQ:Q:DIFF ON

Prior to A.02.00

Input/Output, I/Q, Q Setup

When Differential Input = On, the analyzer will check for attenuation mismatches between the Q and Q-bar ports. If the difference in attenuation values exceeds 0.5 dB an error condition will be set.

54

Common Measurement Functions 1

Couplings

Preset

State Saved

Some active probes include built-in differential capability. When one of these probes is sensed, this key is disabled. Since the differential capability is handled in the probe, the Analyzer will use only the main port and the key will show that the Analyzer's Differential Input mode is Off (indicating that the complementary port not in use).

When a differential probe is not sensed and Q Same as I is On, the value set for I will be copied to Q. This key is disabled when Q Same as I is On.

Off

On

This is unaffected by a Preset but is set to the default value on a "Restore

Input/Output Defaults" or "Restore System Defaults->All"

Off | On Range

Q Input Z

Selects the input impedance for the Q channel. The impedance applies to both the Q and Q-bar ports.

The input impedance controls the hardware signal path impedance match. It is not used for converting voltage to power. The voltage to power conversion always uses the Reference Z parameter. The Reference Z parameter applies to both I and Q channels.

Remote Command

Example

Initial S/W Revision

Key Path

Notes

Couplings

Preset

State Saved

Range

:INPut[1]:IQ:Q:IMPedance LOW|HIGH

:INPut[1]:IQ:Q:IMPedance?

Set the Q channel input impedance to 1 M

Ω

INP:IQ:Q:IMP HIGH

Prior to A.02.00

Input/Output, I/Q, Q Setup

LOW = 50

Ω, HIGH = 1 MΩ

Input impedance is a built-in characteristic of a probe. Therefore, whenever a probe is sensed, this key is disabled and the value is set to match the probe.

When no probe is sensed and Q Same as I is On, the value set for I will also be copied to Q. This key is disabled when Q Same as I is On.

LOW

On

This is unaffected by Preset but is set to the default value on a "Restore

Input/Output Defaults" or "Restore System Defaults->All"

50

Ω | 1 MΩ

Q Skew

Sets the skew factor for the Q channel. The skew will shift the channel's data in time. Use this to compensate for

55

Common Measurement Functions 1

differences in the electrical lengths of the input paths due to cabling and probes.

Key Path

Remote Command

Preset

State Saved

Range

Example

Initial S/W Revision

Input/Output, I/Q, Q Setup

[:SENSe]:CORRection:IQ:Q:SKEW <seconds>

[:SENSe]:CORRection:IQ:Q:SKEW?

0

Yes

This is unaffected by a Preset but is set to the default value on a "Restore

Input/Output Defaults" or "Restore System Defaults->All"

0 s to 100 ns

Delay the data for the Q channel by 10 ns.

CORR:IQ:Q:SKEW 10 ns

Prior to A.02.00

Q Probe

Accesses the probe setup parameters for the Q channel. See “I/Q Probe Setup” on page 56 .

Key Path

State Saved

Readback Text

Input/Output, I/Q, Q Setup

No

[<Q port probe id>]

This is reporting the type of probe sensed on the Q port. There is no parameter for overriding what is sensed.

Prior to A.02.00

Initial S/W Revision

I/Q Probe Setup

The set of I/Q probe setup parameters will change based on the type of probe that is sensed. All probe types have the Attenuation parameter, and all probe types can be calibrated. The remaining parameters are only available for some probe types and will not be shown when not available. The probe type is determined by and reported for only for the I and Q ports, never the I-bar or Q-bar ports. The menu title will be "<ch>: <probe id>", where "<ch>" is either "I" or "Q" and "<probe id>" is the type of probe. For example, for the I Probe setup with an Agilent 1130A probe connected to the I port, the title will be "I:

1130A".

Probe calibration data is stored for each probe type for each channel. When no probe is sensed, the probe type "Unknown" is used, and this is also treated like a probe type with its own calibration data. When a probe is changed, the calibration data for that probe type for that port is restored. An advisory message will be displayed showing the new probe type and the calibration status. The calibration data is stored permanently (survives a power cycle) and is not affected by a Preset or any of the Restore commands.

When the probe has EEPROM identification (most newer Agilent probes have this), the calibration data is stored by probe serial number and port, so if you have two probes of the same type, the correct calibration data will be used for each. For probes that do not have EEPROM identification, the calibration data is stored by probe type and port and the instrument cannot distinguish between different

56

Common Measurement Functions 1

probes of the same type. In all cases (with or without EEPROM identification), the calibration data is port specific, so it will not follow a specific probe from port to port if the probe is moved.

The "Unknown" probe type is used whenever no probe is sensed. When no calibration data exists for

"Unknown" the latest cable calibration data is used (see Section

“I/Q Guided Calibration” on page 101

).

Attenuation

The attenuation is part of the calibration data stored with the probe type and is initially the value that was returned by the last calibration. You can modify this value and any changes will be stored with the calibration data and will survive power cycles and presets. When a probe calibration is performed the attenuation value will be overwritten by the calibration.

Remote Command

Example

Initial S/W Revision

Key Path

Notes

Preset

State Saved

Range

[:SENSe]:CORRection:IQ:I|Q:ATTenuation:RATio <real>

[:SENSe]:CORRection:IQ:I|Q:ATTenuation:RATio?

Set the attenuation for the current I probe to 100.00:1.

CORR:IQ:I:ATT:RAT 100

Prior to A.02.00

Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe

Each probe type has its own attenuation setting. As probes are changed the attenuation value will reflect the new probe's setting. Changing the attenuation affects only the current probe type's setting and leaves all others unchanged.

Each probe type has its own default. The default for the "Unknown" probe type is 1:1.

Saved with probe calibration data. It survives a power cycle and is not affected by a Preset or Restore.

0.001 to 10000

Remote Command:

Range:

Example:

[:SENSe]:CORRection:IQ:I|Q:ATTenuation <rel_ampl>

[:SENSe]:CORRection:IQ:I|Q:ATTenuation?

–60 dB to +80 dB

Set the attenuation for the current I probe type to 100.00:1.

CORR:IQ:I:ATT 20 dB

Prior to A.02.00

Initial S/W Revision:

Offset

Some active probes have DC offset capability. When one of these probes is connected this control will be visible.

The signal is adjusted for the DC offset before entering the analyzer's port. This allows for removal of a DC offset before hitting the analyzer's input port voltage limits. For example, a signal that varies 1 V peak-to-peak with a DC offset equal to the analyzer's max input voltage would exceed the input limits of the analyzer for half its cycle.

57

Common Measurement Functions 1

Removing the DC offset allows the analyzer to correctly process the entire signal.

Remote Command

Example

Initial S/W Revision

Key Path

Notes

Preset

State Saved

Range

:INPut:OFFSet:I|Q <voltage>

:INPut:OFFSet:I|Q?

Remove a DC offset of –0.5 V from the I channel input.

INP:OFF:I –0.5

Prior to A.02.00

Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe

Only some probe types support Offset. For those that do, each probe type has its own Offset setting. As probes are changed the Offset value will reflect the new probe's setting. Changing the Offset affects only the current probe type's setting and leaves all others unchanged.

0 V

Saved with probe calibration data. It survives power cycle and is not affected by Preset or Restore.

–18 V to +18 V

Coupling

Some probe types allow coupling to reject low frequencies. This will filter out the DC component of a signal that is composed of a DC bias plus some AC signal. This control is visible only for probe types that have this capability.

Remote Command

Example

Initial S/W Revision

Key Path

Notes

Preset

State Saved

Range

Readback Text

:INPut:COUPling:I|Q DC|LFR1|LFR2

:INPut:COUPling:I|Q?

Set the probe to low frequency rejection below 1.7 Hz.

INP:COUP:I LFR1

Prior to A.02.00

Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe

Only some probe types support Coupling. For those that do, each probe type has its own Coupling setting. As probes are changed the Coupling value will reflect the new probe's setting. Changing the Coupling affects only the current probe type's setting and leaves all others unchanged.

DC

Saved with probe calibration data. It survives a power cycle and is not affected by a Preset or Restore.

DC | AC 1.7 Hz LFR1 | AC 0.14 Hz LFR2

DC | LFR1 | LFR2

58

Common Measurement Functions 1

DC

Turns off low frequency rejection, allowing signals down to DC.

Key Path

Example

Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, Coupling

Turn off low frequency rejection on the I channel

INP:COUP:I DC

Prior to A.02.00

Initial S/W Revision

LFR1

Turns on low frequency rejection, rejecting signal component lower than 1.7 Hz.

Key Path

Example

Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, Coupling

Turn on low frequency rejection on the I channel for frequencies lower than

1.7 Hz

INP:COUP:I LFR1

Prior to A.02.00

Initial S/W Revision

LFR2

Turns on low frequency rejection, rejecting signal component lower than 0.14 Hz.

Key Path

Example

Initial S/W Revision

Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, Coupling

Turn on low frequency rejection on the I channel for frequencies lower than

0.14 Hz

INP:COUP:I LFR2

Prior to A.02.00

Calibrate

Invokes the guided probe calibration. The guided probe calibration is context sensitive and depends on the channel

(I or Q) and the Differential Input state. The calibration is only performed on the selected channel. When

Differential Input is on, both the probe attached to the main port and the probe attached to the complementary port are calibrated. When Differential Input is off, only the probe attached to the main port is calibrated. See

“I/Q

Guided Calibration” on page 101

.

Key Path

Readback Text

Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, Coupling

The last calibration date, or if no calibration exists, "(empty)".

Last: <cal date>

<cal time>

Example:

Last: 8/22/2007

1:02:49 PM

59

Common Measurement Functions 1

Initial S/W Revision Prior to A.02.00

Clear Calibration

Clears the calibration data for the current port and probe. It does not clear the data for other probe types or other ports. If the sensed probe has EEPROM identification, only the data for that specific probe is cleared. After this command has completed, the probe calibration state will be the same as if no probe calibration had ever been performed for the specified channel and probe. The probe attenuation will be the default value for that probe type and the Cable Calibration frequency response corrections will be used. This command is dependent on the

Differential Input state. When Differential Input is on, both the data for the probe attached to the main port and the data for the probe attached to the complementary port are cleared. When Differential Input is off, only data for the probe attached to the main port is cleared.

Key Path

Remote Command

Example

Initial S/W Revision

Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe

:CALibration:IQ:PROBe:I|Q:CLEar

Clear the calibration data for the I channel and the current probe (with

EEPROM identification) or probe type (without EEPROM identification).

:CAL:IQ:PROBe:I:CLE

Prior to A.02.00

Reference Z

Sets the value of the impedance to be used in converting voltage to power for the I and Q channels. This does not change the hardware's path impedance (see

“I Input Z” on page 51 ).

Key Path

Remote Command

Preset

State Saved

Range

Example

Initial S/W Revision

Input/Output, I/Q

:INPut:IMPedance:REFerence <integer>

:INPut:IMPedance:REFerence?

50

Ω

Yes

This is unaffected by a Preset but is set to the default value on a "Restore

Input/Output Defaults" or "Restore System Defaults->All"

1

Ω to 1 MΩ

Set the I/Q reference impedance to 50

Ω

INP:IMP:REF 50

Prior to A.02.00

I/Q Cable Calibrate…

Invokes the guided cable calibration. The guided cable calibration steps the user through a calibration of all ports (I, I-bar, Q, and Q-bar) using just a cable (no probe attached). See

“I/Q Guided Calibration” on

60

Common Measurement Functions 1

page 101

.

Key Path

Initial S/W Revision

Input/Output, I/Q

Prior to A.02.00

RF Calibrator

Lets you choose a calibrator signal to look at or turns the calibrator "off" (switches back to the selected input). When one of the calibrator signals is selected, the analyzer routes that signal (an internal amplitude reference) to the analyzer, while leaving the main input selection (RF or I/Q) unchanged.

This function presets to OFF on a Mode Preset, which causes the internal circuitry to switch back to the selected input (RF or I/Q).

Remote Command:

Example:

Dependencies:

[:SENSe]:FEED:AREFerence REF50|REF4800|OFF

[:SENSe]:FEED:AREFerence?

FEED:AREF REF50 !selects the 50 MHz amplitude reference as the signal input.

FEED:AREF REF4800 !selects the 4.8 GHz amplitude reference as the signal input

FEED:AREF OFF !turns the calibrator "off" (switches back to the selected input – RF or I/Q)

Selecting an input (RF or I/Q) turns the Calibrator OFF. This is true whether the input is selected by the keys or with the [:SENSe]:FEED command.

The 4.8 GHz internal reference is only available in some models and frequency range options. If the 4.8 GHz reference is not present, the

4.8 GHz

softkey will be blanked, and if the REF4800 parameter is sent, the analyzer will generate an error.

Preset:

State Saved:

Key Path:

Readback:

OFF

Saved in State

Input/Output

Off, 50 MHz, 4.8 GHz

Backwards Compatibility SCPI: For ESA backwards compatibility, the legacy SCPI command

CALibration:SOURce:STATe <boolean> (ESA’s Amptd Ref Out SCPI) will still be supported and mapped as follows:

When CALibration:SOURce:STATe ON is received [SENSe]:FEED:AREF

REF50 will execute

When CALibration:SOURce:STATe OFF is received [SENSe]:FEED:AREF

OFF will execute

When CALibration:SOURce:STATe? is received, 1 will be returned if any of the references is selected and 0 if the Calibrator is "Off"

Initial S/W Revision: Prior to A.02.00

61

Common Measurement Functions 1

50 MHz

Selects the 50 MHz internal reference as the input signal.

Example:

Key Path:

Readback:

Initial S/W Revision:

:FEED:AREF REF50

Input/Output, RF Calibrator

50 MHz

Prior to A.02.00

4.8 GHz

Selects the 4.8 GHz internal reference as the input signal.

Example:

Key Path:

Readback:

Dependencies:

:FEED:AREF REF4800

Input/Output, RF Calibrator

4.8 GHz

The 4.8 GHz internal reference is only available in some models and frequency range options. If the 4.8 GHz reference is not present, the

4.8 GHz

softkey will be blanked, and if the REF4800 parameter is sent, the analyzer will generate an error.

Initial S/W Revision:

Modified at S/W Revision:

Prior to A.02.00

A.03.00

Off

Switches the input back to the selected input (RF or I/Q)

Example:

Key Path:

Readback:

Initial S/W Revision:

:FEED:AREF OFF

Input/Output, RF Calibrator

Off

Prior to A.02.00

External Gain

Compensates for gain or loss in the measurement system outside the spectrum analyzer. The External

Gain is subtracted from the amplitude readout (or the loss is added to the amplitude readout). So, the displayed signal level represents the signal level at the output of the device-under-test, which can be the input of an external device that provides gain or loss.

Entering an External Gain value does not affect the Reference Level, therefore the trace position on screen changes, as do all of values represented by the trace data. Thus, the values of exported trace data, queried trace data, marker amplitudes, trace data used in calculations such as N dB points, trace math, peak threshold, etc., are all affected by External Gain. Changing the External Gain, even on a trace which is not updating, will immediately change all of the above, without new data needing to be taken.

62

Common Measurement Functions 1

NOTE

Changing the External Gain causes the analyzer to immediately stop the current sweep and prepare to begin a new sweep, but the data will not change until the trace data updates, because the offset is applied to the data as it is taken. If a trace is exported with a nonzero External Gain, the exported data will contain the trace data with the offset applied.

In the Spectrum Analyzer mode, a Preamp is the common external device providing gain or loss. In a measurement application mode like GSM or W-CDMA, the gain or loss could be from a BTS (Base

Transceiver Station) or an MS (Mobile Station). So in the Spectrum Analyzer mode MS and BTS would be grayed out and the only choice would be Ext Preamp. Similarly in some of the digital communications applications, Ext Preamp will be grayed out and you would have a choice of MS or

BTS.

Initial S/W Revision

Key Path

Couplings

Prior to A.02.00

Input/Output

The Ext Preamp, MS, and BS keys may be grayed out depending on which measurement is currently selected. If any of the grayed out keys are pressed, or the equivalent SCPI command is sent, an advisory message is generated.

Readback 1-of-N selection | [variable]

Ext Preamp

This function is similar to the reference level offset function. Both affect the displayed signal level. Ref

Lvl Offset is a mathematical offset only, no analyzer configuration is affected. Ext Preamp gain is used when determining the auto-coupled value of the Attenuator. The External Gain value and the Maximum

Mixer Level settings are both part of the automatic setting equation for the RF attenuation setting. (10 dB of Attenuation is added for every 10 dB of External Gain.)

Note that the Ref Lvl Offset and Maximum Mixer Level are described in the Amplitude section. They are reset by the instrument Preset. The External Preamp Gain is reset by the "Restore Input/Output

Defaults" or "Restore System Defaults->All functions. . The External Gain is subtracted from the amplitude readout so that the displayed signal level represents the signal level at the output of the device-under-test, which is the input of the external device that is providing gain or loss.

Remote Command:

Example:

Notes:

[:SENSe]:CORRection:SA[:RF]:GAIN <rel_ampl>

[:SENSe]:CORRection:SA[:RF]:GAIN?

CORR:SA:GAIN 10 !sets the Ext Gain value to 10 dB

CORR:SA:GAIN –10 !sets the Ext Gain value to –10 dB (that is, an attenuation of 10 dB)

Does not auto return.

63

Common Measurement Functions 1

Dependencies:

Preset:

The reference level limits are determined in part by the External Gain/Atten,

Max Mixer Level, and RF Atten.

This key is grayed out in Modes that do not support External Gain

This is unaffected by Preset but is set to 0 dB on a "Restore Input/Output

Defaults" or "Restore System Defaults->All"

State Saved:

Min:

Saved in instrument state

–81.90 dB

Max:

Key Path:

81.90 dB

Readback:

Input/Output, External Gain

Preamp Gain, <Ext Gain value> dB

Backwards Compatibility SCPI: [:SENSe]:CORRection:OFFSet[:MAGNitude]

Initial S/W Revision: Prior to A.02.00

MS

Sets an external gain/attenuation value for MS (Mobile Station) tests.

Remote Command:

Example:

Notes:

Dependencies:

Preset:

[:SENSe]:CORRection:MS[:RF]:GAIN <rel_ampl>

[:SENSe]:CORRection:MS[:RF]:GAIN?

CORR:MS:GAIN 10 !sets the Ext Gain value to 10 dB

CORR:MS:GAIN –10 !sets the Ext Gain value to –10 dB (that is, a loss of 10 dB.)

Does not auto return.

The reference level limits are determined in part by the External Gain, Max

Mixer Level, RF Atten

This key is grayed out in modes that do not support MS.

This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output

Defaults" or "Restore System Defaults->All"

State Saved:

Min:

Max:

Key Path:

Readback:

Saved in instrument state.

–100 dB

100 dB

Input/Output, External Gain

MS, <Ext Gain value> dB

64

Common Measurement Functions 1

Backwards Compatibility SCPI: [:SENSe]:CORRection:MS[:RF]:LOSS

Backwards Compatibility SCPI: [:SENSe]:CORRection:MS[:RF]:LOSS <rel_ampl>

[:SENSe]:CORRection:MS[:RF]:LOSS?

Important notes regarding the alias commands:

A positive value of <rel_ampl> in the above command means a loss and a negative value indicates a gain. So, for example, sending the command a) CORR:MS:LOSS 10 dB will set the value on the softkey and the active function to –10 dB since the softkey and the active function always show the

Gain.

The query CORR:MS:LOSS? returns 10 dB

The query CORR:MS:GAIN? returns –10 dB b) CORR:MS:LOSS –10 dB will set the value on the softkey and the active function to 10 dB since the softkey and the active function always show the

Gain

The query CORR:MS:LOSS? returns –10 dB

The query CORR:MS:GAIN? returns 10 dB

Initial S/W Revision: Prior to A.02.00

BTS

Sets an external attenuation value for BTS (Base Transceiver Station) tests.

Remote Command:

Example:

Notes:

Dependencies:

Preset:

[:SENSe]:CORRection:BTS[:RF]:GAIN <rel_ampl>

[:SENSe]:CORRection:BTS[:RF]:GAIN?

CORR:BTS:GAIN 10 !sets the Ext Gain value to 10 dB

CORR:BTS:GAIN –10 !sets the Ext Gain value to –10 dB (that is, a loss of 10 dB.)

Does not auto return.

The reference level limits are determined in part by the External Gain, Max

Mixer Level, RF Atten

This key is grayed out in modes that do not support BTS.

This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output

Defaults" or "Restore System Defaults->All"

State Saved:

Min:

Max:

Key Path:

Saved in instrument state.

–100 dB

100 dB

Input/Output, External Gain

65

Common Measurement Functions 1

Readback: BTS, <Ext Gain value> dB

Backwards Compatibility SCPI: [:SENSe]:CORRection:BTS[:RF]:LOSS

Backwards Compatibility SCPI: [:SENSe]:CORRection:BTS[:RF]:LOSS <rel_ampl>

[:SENSe]:CORRection:BTS[:RF]:LOSS?

Important notes regarding the alias commands:

A positive value of <rel_ampl> in the above command means a loss and a negative value indicates a gain. So, for example, sending the command a) CORR:BTS:LOSS 10 dB will set the value on the softkey and the active function to –10 dB since the softkey and the active function always show the

Gain.

The query CORR:BTS:LOSS? returns 10 dB

The query CORR:BTS:GAIN? returns –10 dB b) CORR:BTS:LOSS –10 dB will set the value on the softkey and the active function to 10 dB since the softkey and the active function always show the

Gain

The query CORR:BTS:LOSS? returns –10 dB

The query CORR:BTS:GAIN? returns 10 dB

Initial S/W Revision: Prior to A.02.00

I Ext Gain

This function affects only the I channel input, except when the Input Path is I+jQ. In I+jQ this setting is applied to both I and Q channel inputs. It is not available unless the Baseband I/Q option (BBA) is installed.

Remote Command

Example

Initial S/W Revision

Key Path

Notes

Preset

[:SENSe]:CORRection:IQ:I:GAIN <rel_ampl>

[:SENSe]:CORRection:IQ:I:GAIN?

Set the I Ext Gain to 10 dB

CORR:IQ:I:GAIN 10

Set the I Ext Gain to –10 dB (that is, a loss of 10 dB.)

CORR:IQ:I:GAIN –10

Prior to A.02.00

Input/Output, External Gain

Not available unless option BBA is installed

0 dB

This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output

Defaults" or "Restore System Defaults->All"

66

Common Measurement Functions 1

State Saved

Min

Max

Readback Text

Saved in instrument state.

–100 dB

100 dB

I Gain, <I Ext Gain> dB

Q Ext Gain

This function affects only the Q channel input and only when the Input Path is not I+jQ. It is not available unless the Baseband I/Q option (BBA) is installed.

Remote Command

Example

Initial S/W Revision

Key Path

Notes

Preset

State Saved

Min

Max

Readback Text

[:SENSe]:CORRection:IQ:Q:GAIN <rel_ampl>

[:SENSe]:CORRection:IQ:Q:GAIN?

Set the Q Ext Gain to 10 dB

CORR:IQ:Q:GAIN 10

Set the Q Ext Gain to –10 dB (that is, a loss of 10 dB.)

CORR:IQ:Q:GAIN –10

Prior to A.02.00

Input/Output, External Gain

Not available unless option BBA is installed.

0 dB

This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output

Defaults" or "Restore System Defaults->All"

Saved in instrument state.

–100 dB

100 dB

Q Gain, <I Ext Gain> dB

Restore Input/Output Defaults

This selection causes the group of settings and data associated with the

Input/Output

key to be a reset to their default values. This level of Restore System Defaults does not affect any other system settings or mode settings and does not cause a mode switch. All the features described in this section are reset using this key, including Input Corrections and Data (described in the Corrections section).

Example:

Notes:

:SYST:DEF INP !presets all the Input/Output variables to their factory default values.

Please refer to the Utility Functions for information about Restore System

Defaults and the complete description of the :SYSTem:DEFault INPut: command.

67

Common Measurement Functions 1

Key Path:

Initial S/W Revision:

Input/Output

Prior to A.02.00

Data Source

Gives you the choice of either using a hardware input signal as the input or raw data stored in a data storage buffer from an earlier acquisition. You can also share raw data across certain measurements that support this feature. The measurements must be capable of storing raw data. There are three choices under this menu. You can select "Inputs" which is the same as selecting one of the inputs from the input port, for example RF, AREF, I/Q, or IFALign. Selecting "Capture Buffer" allows you to use data that has been stored earlier in the same measurement or from a previous measurement using the "Current Meas

-> Capture Buffer" feature. Selecting "Recorded Data" allows you to playback long data capture records stored in the record buffer.

When you make a recording (see “Record Data Now” on page 70 ) or when you recall a recording (see

the Recall section) the data source is automatically set to Recorded Data. You can toggle the data source between Inputs and the current Recording (if there is one). That is, the recording remains in memory until it is replaced by a new recording, or the application is closed.

Remote Command:

Example:

Dependencies:

[:SENSe]:FEED:DATA INPut|STORed|RECorded

[:SENSe]:FEED:DATA?

FEED:DATA REC

FEED:DATA?

Not all inputs are available in all modes. Unavailable keys are grayed out.

Notes:

Preset:

State Saved:

Key Path:

Readback:

68

INPuts = Inputs

STORed = Capture Buffer

RECorded = Record Data Buffer

This is unaffected by Preset but is set to INPut on a "Restore Input/Output

Defaults" or "Restore System Defaults->All"

Saved in state

Input/Output

Variable

Common Measurement Functions 1

Backwards Compatibility SCPI: [:SENSe]:FEED:SOURce INPut|STORed

[:SENSe]:FEED:SOURce?

Initial S/W Revision: Prior to A.02.00

Inputs

Sets the measurement to use the input selections (RF, AREF, I/Q)

Example: FEED:DATA INP !causes the measurement to look at the input selection

Notes:

Key Path:

Readback:

Initial S/W Revision:

Does not auto return.

Input/Output, Data Source

Inputs

Prior to A.02.00

Capture Buffer

Some WCDMA and demod measurements support this feature. This allows sharing of the raw data across certain measurements. If you want to make another measurement on the same signal, you would store that raw data using the "Current Meas -> Capture Buffer" key. Then the data is available for the next measurement to use. You must have raw data stored in the instrument memory before the Capture

Buffer choice is available for use.

If you switch to a measurement that does not support this feature, then the instrument switches to use

"Inputs" and grays out this key. If the grayed out key is pressed, it generates a message.

Example:

Notes:

Dependencies:

Key Path:

Readback:

Initial S/W Revision:

FEED:DATA STOR !causes stored measurement data to be used with a different measurement that supports this.

Does not auto return. This key is grayed out when you switch to a measurement that does not support this feature.

Grayed out in the SA measurement.

Input/Output, Data Source

Stored Data

Prior to A.02.00

Recorded Data

Directs the instrument to get data from the record data buffer in the measurement, rather than from the

RF Input Signal.

Example:

Notes:

FEED:DATA REC !causes the measurement to extract data from the record data buffer.

Does not auto return.

69

Common Measurement Functions 1

Dependencies:

Key Path:

Readback:

Initial S/W Revision:

Grayed out in the SA measurement.

Input/Output, Data Source

Recorded Data

Prior to A.02.00

Current Meas -> Capture Buffer

Pressing this key stores the raw data of one measurement in the internal memory of the instrument where it can then be used by a different measurement by pressing "Stored Data". When raw data is stored, then the data source selection switch automatically changes to "Stored Data". Stored raw data cannot be directly accessed by a user. There is no save/recall function to save the raw data in an external media.

However if you want to get the stored raw data, you must first perform a measurement using the stored raw data. Now you can access the used raw data, which is the same as stored raw data, using the FETch or READ commands.

Remote Command:

Example:

Dependencies:

Notes:

[:SENSe]:FEED:DATA:STORe

FEED:DATA:STOR !stores recorded data

Grayed out in the SA measurement.

This is command only, there is no query

Key Path:

Input/Output, Data Source

Backwards Compatibility SCPI: [:SENSe]:FEED:SOURce:STORe

Initial S/W Revision: Prior to A.02.00

Record Data Now

This causes the data source to change to Inputs (if it is not already set) and a recording is made with the current instrument setup. The length of the recording must be specified in advance.

This key changes to

Abort Recording

once the recording process has started. It changes back when the recording is complete.

70

The following dialogs show the progress of the recording:

Common Measurement Functions 1

This key is also available in the Sweep/Control menu.

Remote Command

Example

Initial S/W Revision

Key Path

Mode

Notes

Dependencies

Couplings

[:SENSe]:RECording:INITiate[:IMMediate]

REC:INIT

Prior to A.02.00

Input/Output, Data Source

VSA

This is command only, there is no query. See the Recall functionality to access previously saved data.

Grayed out in the SA measurement.

Changes Data source to Recorded Data.

Remote Command:

Example:

Key Path:

Notes:

[:SENSe]:RECording:ABORt

REC:ABOR

Input/Output, Data Source

This is command only, there is no query. The command does nothing if it is sent when there is no recording in progress.

Prior to A.02.00

Initial S/W Revision:

Record Length

This specifies the length of the next recording. (You cannot use this to modify the length of the current recording.) The length defaults to seconds, but you can also specify it in points at the current sample rate,

71

Common Measurement Functions 1

or in time records at the current time record length.

Remote Command

Example

Initial S/W Revision

Key Path

Mode

Notes

Preset

State Saved

Min

Max

Readback

[:SENSe]:RECording:LENGth <real>,SEConds|RECords|POINts

[:SENSe]:RECording:LENGth:STATe MAX|MANual

[:SENSe]:RECording:LENGth:STATe?

REC:LENG 20,REC

REC:LENG 4.1E–4,SEC

REC:LENG:STAT MAX

REC:LENG:STAT?

Prior to A.02.00

Input/Output, Data Source

VSA

There is no default unit. The unit must be specified.

The length command does not have a query form. Length information is queried using the two commands following this table.

If set to MAX, all of the available "recording memory" us used.

50 Records, Manual

No

0

Depends on memory available.

<value><Seconds|Points|Records>

Remote Command

Example

Initial S/W Revision

Mode

Notes

[:SENSe]:RECording:LENGth:VALue?

REC:LENG:VAL?

Prior to A.02.00

VSA

Query Only

Returns the first (numeric) parameter of the most recent

[:SENSe]:RECording:LENGth command.

50 Records Preset

Remote Command

Example

Initial S/W Revision

[:SENSe]:RECording:LENGth:UNIT?

REC:LENG:UNIT?

Prior to A.02.00

72

Common Measurement Functions 1

Mode

Notes

VSA

Query Only

Returns the second parameter of the most recent

[:SENSe]:RECording:LENGth command. Possible values are

SEC|REC|POIN. If no second parameter was sent, then the return value is

SEC.

RECords Preset

Corrections

This key accesses the Amplitude Corrections menu.

Amplitude Corrections arrays can be entered, sent over SCPI, or loaded from a file. They allow you to correct the response of the analyzer for various use cases. The X-series supports four separate

Corrections arrays, each of which can contain up to 2000 points. They can be turned on and off individually and any or all can be on at the same time.

Trace data is in absolute units and corrections data is in relative units, but we want to be able to display trace data at the same time as corrections data. Therefore we establish a reference line to be used while building or editing a Corrections table. The reference line is halfway up the display and represents 0 dB of correction. It is labeled “0 dB CORREC”. It is drawn in blue.

Corrections data is always in dB. Whatever dB value appears in the correction table represents the correction to be applied to that trace at that frequency. So if a table entry shows 30 dB that means we

ADD 30 dB to each trace to correct it before displaying it.

In zero span, where the frequency is always the center frequency of the analyzer, we apply the

(interpolated) correction for the center frequency to all points in the trace. In the event where there are two correction amplitudes at the center frequency, we apply the first one in the table.

Note that the corrections are applied as the data is taken; therefore, a trace in

View

(Update Off) will not be affected by changes made to the corrections table after the trace is put in

View

.

Initial S/W Revision

Key Path

Mode

Dependencies

Preset

A.02.00

Input/Output, Corrections

SA, DVB-T/H, DTMB, SEQAN

This key will only appear if you have the proper option installed in your instrument.

Amplitude correction may not be available in all modes; if a mode does not support amplitude correction, the Corrections key should be blanked while in that mode. If an application supports corrections but the current measurement does not, then the key should be grayed out in that measurement

Corrections arrays are reset (deleted) by Restore Input/Output Defaults.

They survive shutdown and restarting of the analyzer application, which means they will survive a power cycle.

73

Common Measurement Functions 1

Select Correction

Specifies the selected correction. The term "selected correction" is used throughout this document to specify which correction will be affected by the functions.

Initial S/W Revision

Key Path

Mode

Notes

Preset

Readback

A.02.00

Input/Output, Corrections

SA

The selected correction is remembered even when not in the correction menu.

Set to Correction 1 by Restore Input/Output Defaults

Correction 1|Correction 2|Correction 3|Correction 4|Correction 5|Correction

6

Correction On/Off

Turning the Selected Correction on allows the values in it to be applied to the data. This also automatically turns on "Apply Corrections" (sets it to ON), otherwise the correction would not take effect.

A new sweep is initiated if an amplitude correction is switched on or off. Note that changing, sending or loading corrections data does NOT directly initiate a sweep, however in general these operations will turn corrections on, which DOES initiate a sweep.

Remote Command:

Example:

Dependencies:

Preset:

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6[:STATe]

ON|OFF|1|0

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6[:STATe]?

SENS:CORR:CSET1 ON

Turning this on automatically turns on "Apply Corrections"

Only the first correction array (Correction 1) supports antenna units. When this array is turned on, and it contains an Antenna Unit other than “None”, the

Y Axis Unit of the analyzer is forced to that Antenna Unit. All other Y Axis

Unit choices are grayed out.

Note that this means that a correction file with an Antenna Unit can only be loaded into the Corrections 1 register. Consequently only for Correction 1 does the dropdown in the Recall dialog include.ant, and if an attempt is made to load a correction file into any other Correction register which DOES contain an antenna unit, a Mass Storage error is generated.

This command will generate an “Option not available” error unless you have the proper option installed in your instrument.

Not affected by a Preset. Set to OFF by

Restore Input/Output Defaults

State Saved:

Key Path:

Saved in instrument state.

Input/Output, Corrections

74

Common Measurement Functions 1

Initial S/W Revision: A.02.00

Properties

Accesses a menu that lets you set the properties of the selected correction.

Key Path

Initial S/W Revision

Input/Output, Corrections

A.02.00

Select Correction

Specifies the selected correction. The term "selected correction" is used throughout this document to specify which correction will be affected by the functions.

Initial S/W Revision

Key Path

Notes

Preset

Readback

A.02.00

Input/Output, Corrections, Properties

The selected correction is remembered even when not in the correction menu.

Set to Correction 1 by Restore Input/Output Defaults.

Correction 1|Correction 2|Correction 3|Correction 4|Correction 5|Correction

6

Antenna Unit

For devices (like antennae) which make measurements of field strength or flux density, the correction array should contain within its values the appropriate conversion factors such that, when the data on the analyzer is presented in dB

μV, the display is calibrated in the appropriate units. The "Antenna Unit" used for the conversion is contained within the corrections array database. It may be specified by the user or loaded in from an external file or SCPI.

When an array with an Antenna Unit other than "None" is turned on, the Y Axis Unit of the analyzer is forced to that unit. When this array is turned on, and it contains an Antenna Unit other than “None”, the Y Axis Unit of the analyzer is forced to that Antenna Unit., and all other Y Axis Unit choices are grayed out.

Antenna Unit does not appear in all Modes that support Corrections. Only the modes listed in the Mode row of the table below support Antenna Units.

Remote Command

Example

Initial S/W Revision

Key Path

Mode

[:SENSe]:CORRection:CSET[1]|2|3|4:ANTenna[:UNIT]

GAUSs|PTESla|UVM|UAM|NOConversion

[:SENSe]:CORRection:CSET[1]|2|3|4:ANTenna[:UNIT]?

CORR:CSET:ANT GAUS

A.02.00

Input/Output, Corrections, Properties

SA

75

Common Measurement Functions 1

Dependencies

Preset

Only the first correction array (Correction 1) supports antenna units.

Note that this means that a correction file with an Antenna Unit can only be loaded into the Corrections 1 register. Consequently only for Correction 1 does the dropdown in the Recall dialog include.ant, and if an attempt is made to load a correction file into any other Correction register which DOES contain an antenna unit, a Mass Storage error is generated.

Unaffected by Preset. Set to NOC by Restore Input/Output Defaults

State Saved Saved in State

dBµV/m

Sets the antenna unit to dBµV/m. If this correction is turned on, and Apply Corrections is on, the Y Axis Unit will then be forced to dBµV/m and all other Y Axis Unit selections will be grayed out.

Example:

Key Path:

:CORR:CSET2:ANT UVM

Input/Output, Corrections, Properties, Antenna Unit

Readback:

Initial S/W Revision:

"dBµV/m"

A.02.00

dBµA/m

Sets the antenna unit to dBµA/m. If this correction is turned on, and Apply Corrections is on, the Y Axis Unit will then be forced to dBµA/m and all other Y Axis Unit selections will be grayed out.

Example:

Key Path:

:CORR:CSET2:ANT UVA

Input/Output, Corrections, Properties, Antenna Unit

Readback:

Initial S/W Revision:

" dBµA/m"

A.02.00

dBpT

Sets the antenna unit to dBpT. If this correction is turned on, and Apply Corrections is on, the Y Axis Unit will then be forced to dBpT and all other Y Axis Unit selections will be grayed out.

Example:

Key Path:

:CORR:CSET3:ANT PTES

Input/Output, Corrections, Properties, Antenna Unit

Readback:

Initial S/W Revision:

"dBpT"

A.02.00

76

Common Measurement Functions 1 dBG

Sets the antenna unit to dBG. If this correction is turned on, and Apply Corrections is on, the Y Axis Unit will then be forced to dBG and all other Y Axis Unit selections will be grayed out.

Example:

Key Path:

:CORR:CSET:ANT GAUS

Input/Output, Corrections, Properties, Antenna Unit

Readback:

Initial S/W Revision:

" dBG"

A.02.00

None

Selects no antenna unit for this Correction set. Thus no Y Axis unit will be forced.

Example:

Key Path:

:CORR:CSET4:ANT NOC

Input/Output, Corrections, Properties, Antenna Unit

Readback:

Initial S/W Revision:

"None"

A.02.00

Frequency Interpolation

This setting controls how the correction values per-bucket are calculated. We interpolate between frequencies in either the logarithmic or linear scale.

This setting is handled and stored individually per correction set.

See

“Interpolation” on page 77

Remote Command:

Example:

Preset:

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:X:SPACing

LINear|LOGarithmic

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:X:SPACing?

CORR:CSET:X:SPAC LIN

Unaffected by a Preset. Set to Linear by Restore Input/Output Defaults.

State Saved:

Key Path:

Initial S/W Revision:

Saved in instrument state.

Input/Output, Corrections, Properties

A.02.00

Interpolation

For each bucket processed by the application, all of the correction factors at the frequency of interest (center frequency of each bucket) are summed and added to the amplitude. All trace operations and post processing treat this post-summation value as the true signal to use.

77

Common Measurement Functions 1

To effect this correction, the goal, for any particular start and stop frequency, is to build a correction trace, whose number of points matches the current Sweep Points setting of the instrument, which will be used to apply corrections on a bucket by bucket basis to the data traces.

For amplitudes that lie between two user specified frequency points, we interpolate to determine the amplitude value. You may select either linear or logarithmic interpolation between the frequencies.

If we interpolate on a log scale, we assume that the line between the two points is a straight line on the log scale.

For example, let’s say the two points are (2,4) and (20,1). A straight line between them on a log scale looks like:

On a linear scale (like that of the spectrum analyzer), this translates to:

On the other hand, if we interpolate on a linear scale, we assume that the two points are connected by a straight line on the linear scale, as below:

78

Common Measurement Functions 1

The correction to be used for each bucket is taken from the interpolated correction curve at the center of the bucket.

Description

Sets an ASCII description field which will be stored in an exported file. Can be displayed in the active function area by selecting as the active function, if desired to be in a screen dump.

Remote Command:

Example:

Notes:

Preset:

State Saved:

Key Path:

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DESCription

"text"

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DESCription?

:CORR:CSET1:DESC "11941A Antenna correction"

45 chars max; may not fit on display if max chars used

Unaffected by a Preset. Set to empty by

Restore Input/Output Defaults

Saved in instrument state.

Input/Output, Corrections, Properties

Initial S/W Revision: A.02.00

Comment

Sets an ASCII comment field which will be stored in an exported file. Can be displayed in the active function area by selecting as the active function, if desired to be in a screen dump.

Remote Command:

Example:

Notes:

Preset:

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:COMMent "text"

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:COMMent?

:CORR:CSET1:COMM "this is a comment"

45 chars max; may not fit on display if max chars used

Unaffected by Preset. Set to empty by

Restore Input/Output Defaults

79

Common Measurement Functions 1

State Saved:

Key Path:

Saved in State

Input/Output, Corrections, Properties

Initial S/W Revision: A.02.00

Edit

Invokes the integrated editing facility for this correction set.

When entering the menu, the editor window turns on, the selected correction is turned

On

,

Apply

Corrections

is set to

On

, the amplitude scale is set to

Log

, and the Amplitude Correction (“Ampcor”) trace is displayed. The actual, interpolated correction trace is shown in green for the selected correction.

Note that since the actual interpolated correction is shown, the correction trace may have some curvature to it. This trace represents only the correction currently being edited, rather than the total, accumulated amplitude correction for all amplitude corrections which are currently on, although the total, accumulated correction for all corrections which are turned on is still applied to the data traces.

Because corrections data is always in dB, but the Y-axis of the analyzer is in absolute units, it is necessary to establish a reference line for display of the Corrections data. The reference line is halfway up the display and represents 0 dB of correction. It is labeled “0 dB CORREC”. It is drawn in blue.

Corrections data is always in dB. Whatever dB value appears in the correction table represents the correction to be applied to that trace at that frequency. So if a table entry shows 30 dB that means we

ADD 30 dB to each trace to correct it before displaying it. By definition all points are connected. If a gap is desired for corrections data, enter 0 dB.

Note that a well-designed Corrections array should start at 0 dB and end at 0 dB. This is because whatever the high end point is will be extended to the top frequency of the instrument, and whatever the low end point is will be extended down to 0 Hz. So for a Corrections array to have no effect outside its range, you should start and end the array at 0 dB.

NOTE

The table editor will only operate properly if the analyzer is sweeping, because its updates are tied to the sweep system. Thus, you should not try to use the editor in single sweep, and it will be sluggish during compute-intensive operations like narrow-span FFT sweeps.

When exiting the edit menu (by using the

Return

key or by pressing an instrument front-panel key), the editor window turns off and the Ampcor trace is no longer displayed; however,

Apply Corrections

remains

On

, any correction that was on while in the editor remains on, and the amplitude scale returns to its previous setting.

Corrections arrays are not affected by a Preset, because they are in the Input/Output system. They also survive shutdown and restarting of the analyzer application, which means they will survive a power cycle.

Key Path

Initial S/W Revision

Input/Output, Corrections

A.02.00

80

Common Measurement Functions 1

Navigate

Lets you move through the table to edit the desired point.

Initial S/W Revision

Key Path

Notes

Min

Max

A.02.00

Input/Output, Corrections, Edit

There is no value readback on the key

1

2000

Frequency

Lets you edit the frequency of the current row.

Initial S/W Revision

Key Path

Notes

Min

Max

A.02.00

Input/Output, Corrections, Edit

There is no value readback on the key.

0

1 THz

Amplitude

Lets you edit the Amplitude of the current row.

Initial S/W Revision

Key Path

Notes

Min

Max

A.02.00

Input/Output, Corrections, Edit

There is no value readback on the key.

–1000 dB

1000 dB

Insert Point Below

Inserts a point below the current point. The new point is a copy of the current point and becomes the current point.

The new point is not yet entered into the underlying table, and the data in the row is displayed in light gray.

Key Path

Initial S/W Revision

Input/Output, Corrections, Edit

A.02.00

Delete Point

Deletes the currently-selected point, whether or not that point is being edited, and selects the Navigate functionality. The point following the currently-selected point (or the point preceding if there is none) will be

81

Common Measurement Functions 1

selected.

Key Path

Initial S/W Revision

Input/Output, Corrections, Edit

A.02.00

Scale X Axis

Matches the X Axis to the selected Correction, as well as possible. Sets the Start and Stop Frequency to contain the minimum and maximum Frequency of the selected Correction. The range between Start Frequency and Stop

Frequency is 12.5% above the range between the minimum and maximum Frequency, so that span exceeds this range by one graticule division on either side. If in zero-span, or there is no data in the Ampcor table, or the frequency range represented by the table is zero, no action is taken. Standard clipping rules apply if the value in the table is outside the allowable range for the X axis.

Key Path

Initial S/W Revision

Input/Output, Corrections, Edit

A.02.00

Delete Correction

Deletes the correction values for this set. When this key is pressed a prompt is placed on the screen that says “Please press Enter or OK key to delete correction. Press ESC or Cancel to close this dialog.” The deletion is only performed if you press OK or Enter; if so, after the deletion, the informational message

“Correction deleted” appears in the MSG line.

Remote Command:

Example:

Notes:

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DELete

CORR:CSET:DEL

CORR:CSET1:DEL

CORR:CSET4:DEL

Pressing this key when no corrections are present is accepted without error.

Key Path:

Input/Output, Corrections

Initial S/W Revision: A.02.00

Apply Corrections

Applies amplitude corrections which are marked as ON to the measured data. If this is set to OFF, then no amplitude correction sets will be used, regardless of their individual on/off settings. If set to ON, the

corrections that are marked as ON (see “Correction On/Off” on page 74

) are used.

Remote Command:

Example:

[:SENSe]:CORRection:CSET:ALL[:STATe] ON|OFF|1|0

[:SENSe]:CORRection:CSET:ALL[:STATe]?

SENS:CORR:CSET:ALL OFF

This command makes sure that no amplitude corrections are applied, regardless of their individual on/off settings.

82

Common Measurement Functions 1

Preset: Not affected by Preset. Set to OFF by Restore Input/Output Defaults

State Saved:

Key Path:

Initial S/W Revision:

Saved in instrument state.

Input/Output, Corrections

A.02.00

Delete All Corrections

Erases all correction values for all 4 Amplitude Correction sets.

When this key is pressed a prompt is placed on the screen that says “Please press Enter or OK key to delete all corrections. Press ESC or Cancel to close this dialog.” The deletion is only performed if you press OK or Enter; if so, after the deletion, the informational message “All Corrections deleted” appears in the MSG line.

Remote Command:

Example:

Key Path:

Initial S/W Revision:

[:SENSe]:CORRection:CSET:ALL:DELete

CORR:CSET:ALL:DEL

Input/Output, Corrections

A.02.00

Remote Correction Data Set Commands

Set (Replace) Data (Remote Command Only)

The command takes an ASCII series of alternating frequency and amplitude points, each value separated by commas.

The values sent in the command will totally replace all existing correction points in the specified set.

An Ampcor array can contain 2000 points maximum.

Remote Command:

Example:

Preset:

State Saved:

Min:

Max:

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DATA <freq>,

<ampl>, . . .

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DATA?

CORR:CSET1:DATA 10000000,–1.0,20000000,1.0

This defines two correction points at (10 MHz, –1.0 dB) and (20 MHz, 1.0 dB) for correction set 1.

Empty after Restore Input/Output Defaults. Survives a shutdown or restart of analyzer application (including a power cycle).

Saved in instrument state.

Freq: 0 Hz

Amptd: –1000 dBm

Freq: 1 THz

Amptd: +1000 dBm

83

Common Measurement Functions 1

Initial S/W Revision: A.02.00

Merge Correction Data (Remote Command Only)

The command takes an ASCII series of alternating frequency and amplitude points, each value separated by commas. The difference between this command and Set Data is that this merges new correction points into an existing set.

Any new point with the same frequency as an existing correction point will replace the existing point’s amplitude with that of the new point.

An Ampcor array can contain 2000 total points, maximum.

Remote Command:

Example:

Preset:

Min:

Max:

[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DATA:MERGe

<freq>, <ampl>, ...

CORR:CSET1:DATA:MERGE 15000000,–5.0,25000000,5.0

This adds two correction points at (15 MHz, –5.0 dB) and (25 MHz, 5.0 dB) to whatever values already exist in correction set 1.

Empty after Restore Input/Output Defaults. Survives shutdown/restart of analyzer application (including power cycle)

Freq: 0 Hz

Amptd: –1000 dBm

Freq: 1 THz

Amptd: +1000 dBm

A.02.00

Initial S/W Revision:

Freq Ref In

Specifies the frequency reference as being the internal reference, external reference or sensing the presence of an external reference.

When the frequency reference is set to internal, the internal 10 MHz reference is used even if an external reference is connected.

When the frequency reference is set to external, the instrument will use the external reference. However, if there is no external signal present, or it is not within the proper amplitude range, an error condition detected message is generated. When the external signal becomes valid, the error is cleared.

If Sense is selected, the instrument checks whether a signal is present at the external reference connector and will automatically switch to the external reference when a signal is detected. When no signal is present, it automatically switches to the internal reference. No message is generated as the reference switches between external and internal. The monitoring of the external reference occurs approximately on 1 millisecond intervals, and never occurs in the middle of a measurement acquisition, only at the end of the measurement (end of the request).

If for any reason the instrument’s frequency reference is not able to obtain lock, Status bit 2 in the

Questionable Frequency register will be true and an error condition detected message is generated.

When lock is regained, Status bit 2 in the Questionable Frequency register will be cleared and an error message is cleared will be sent.

84

Common Measurement Functions 1

If an external frequency reference is being used, you must enter the frequency of the external reference if it is not exactly 10 MHz. The External Ref Freq key is provided for this purpose.

Remote Command:

Preset:

[:SENSe]:ROSCillator:SOURce:TYPE

INTernal|EXTernal|SENSe

[:SENSe]:ROSCillator:SOURce:TYPE?

This is unaffected by a Preset but is set to SENSe on a "Restore Input/Output

Defaults" or "Restore System Defaults->All".

State Saved:

Key Path:

Status Bits/OPC dependencies:

Saved in instrument state.

Input/Output

STATus:QUEStionable:FREQuency bit 2 set if unlocked.

Initial S/W Revision: Prior to A.02.00

Remote Command:

Notes:

[:SENSe]:ROSCillator:SOURce?

The query [SENSe]:ROSCillator:SOURce? returns the current switch setting.

This means:

1. If it was set to SENSe but there is no external reference so the instrument is actually using the internal reference, then this query returns INTernal and not SENSe.

2. If it was set to SENSe and there is an external reference present, the query returns EXTernal and not SENSe.

3. If it was set to EXTernal, then the query returns "EXTernal"

4. If it was set to INTernal, then the query returns INTernal

Preset: SENSe

Backwards Compatibility SCPI: The query [:SENSe]:ROSCillator:SOURce? was a query-only command in

ESA which always returned whichever reference the instrument was using.

The instrument automatically switched to the ext ref if it was present.

In PSA (which had no sensing) the command [:SENSe]:ROSCillator:SOURce set the reference (INT or EXT), so again its query returned the actual routing.

Thus the query form of this command is 100% backwards compatible with both instruments.

Initial S/W Revision:

Remote Command:

Prior to A.02.00

[:SENSe]:ROSCillator:SOURce INTernal|EXTernal

85

Common Measurement Functions 1

Notes: ([:SENSe]:ROSCillator:SOURce:TYPE INTernal|EXTernal and directly sets the routing to either internal or external.)

Initial S/W Revision: Prior to A.02.00

Sense

The external reference is used if a valid signal is sensed at the Ext Ref input. Otherwise the internal reference is used.

Example:

Key Path:

Readback:

Initial S/W Revision:

:ROSC:SOUR:TYPE SENS

Input/Output, Freq Ref In

Sense

Prior to A.02.00

Internal

The internal reference is used.

Example:

Key Path:

Readback:

Initial S/W Revision:

:ROSC:SOUR:TYPE INT

Input/Output, Freq Ref In

Internal

Prior to A.02.00

External

The external reference is used.

Example:

Key Path:

Readback:

Initial S/W Revision:

:ROSC:SOUR:TYPE EXT

Input/Output, Freq Ref In

External

Prior to A.02.00

Ext Ref Freq

This key tells the analyzer the frequency of the external reference. When the external reference is in use

(either because the reference has been switched to External or because the Reference has been switched to Sense and there is a valid external reference present) this information is used by the analyzer to determine the internal settings needed to lock to that particular external reference signal.

For the instrument to stay locked, the value entered must be within 5 ppm of the actual external reference frequency. So it is important to get it close, or you risk an unlock condition.

Note that this value only affects the instrument’s ability to lock. It does not affect any calculations or

86

Common Measurement Functions 1

measurement results. See "Freq Offset" in the Frequency section for information on how to offset frequency values.

Remote Command:

Example:

Notes:

Preset:

Min:

Max:

Key Path:

Default Unit:

Initial S/W Revision:

[:SENSe]:ROSCillator:EXTernal:FREQuency <freq>

[:SENSe]:ROSCillator:EXTernal:FREQuency?

ROSC:EXT:FREQ 20 MHz !sets the external reference frequency to 20 MHz, but does not select the external reference.

ROSC:SOUR:TYPE EXT !selects the external reference.

Still available with Internal selected, to allow setup for when External is in use.

This is unaffected by a Preset but is set to 10 MHz on a "Restore Input/Output

Defaults" or "Restore System Defaults->All"

CXA: 10 MHz

EXA: 10 MHz

MXA: 1 MHz

PXA: 1 MHz

CXA: 10 MHz

EXA: 10 MHz

MXA: 50 MHz

PXA: 50 MHz

Input/Output, Freq Ref In

Hz

Prior to A.02.00

External Reference Lock BW

This control lets you adjust the External Reference phase lock bandwidth. This control is available in some models of the X-Series.

The PXA variable reference loop bandwidth allows an external reference to be used and have the analyzer close-in phase noise improved to match that of the reference. This could result in an improvement of tens of decibels. The choice of “Wide” or “Narrow” affects the phase noise at low offset frequencies, especially 4 to 400 Hz offset. When using an external reference with superior phase noise, we recommend setting the external reference phase-locked-loop bandwidth to wide (60 Hz), to take advantage of that superior performance. When using an external reference with inferior phase noise performance, we recommend setting that bandwidth to narrow (15 Hz). In these relationships, inferior and superior phase noise are with respect to 134 dBc/Hz at 30 Hz offset from a 10 MHz reference.

Because most reference sources have phase noise behavior that falls off at a rate of 30 dB/decade, this is usually equivalent to 120 dBc/Hz at 10 Hz offset.

Remote Command

[:SENSe]:ROSCillator:BANDwidth WIDE|NARRow

[:SENSe]:ROSCillator:BANDwidth?

87

Common Measurement Functions 1

Example

Initial S/W Revision

Key Path

Scope

Dependencies

Preset

ROSC:BAND WIDE

A.04.00

Input/Output, Freq Ref In

Mode Global

This key only appears in analyzers equipped with the required hardware.

This is unaffected by a Preset but is set to Narrow on a "Restore Input/Output

Defaults" or "Restore System Defaults -> All"

Saved in Input/Output state.

State Saved

External Ref Coupling

Only appears with option ERC installed and licensed.

This function lets you couple the sweep system of the analyzer to the state of the External Reference. If

Normal

is selected, data acquisition proceeds regardless of the state of the External Reference. When you select

Ext Ref Out Of Range Stops Acquisition

, the data acquisition (sweep or measurement) stops when either the "521, External ref out of range" or the "503, Frequency Reference unlocked” error is asserted. Note that this will only take place if the

Freq Ref In

selection is

External

.

With the acquisition stopped, the data display will stop updating (even if this occurs in the middle of a sweep or measurement) and no data will be returned to a READ? or MEASure? query; that is, these queries will not complete because the analyzer will not respond to them. Furthermore, no response will be generated to a *WAI? or *OPC? query.

Proper SCPI sequences are shown below, which will always fail to return if the acquisition stops during the requested sweep or measurement. Note that, for predictable operation of this function, it is best to operate the analyzer in single measurement mode (INIT:CONT OFF), because if operating in continuous mode, the analyzer may respond to the above queries even after the acquisition stops, with data left over from the previous acquisition.

:INIT:CONT OFF

:INIT:IMM;*OPC?

--

:INIT:CONT OFF

:INIT:IMM;*WAI?

--

:INIT:CONT OFF

:READ?

--

:INIT:CONT OFF

:MEASure?

When the acquisition ceases, in addition to the error condition(s) described above, an error message will

88

Common Measurement Functions 1

be generated informing you that the acquisition has ceased due to an invalid external reference. This message will stay on the screen while the acquisition is suspended.

If you press the Restart key this message will be taken off the screen and a new acquisition will be attempted; if the External Reference problem persists the message will go right back up. You can also take the message down by changing back to the

Normal

setting of Sweep/Ext Ref Coupling, or by pressing

Freq Ref In, Internal

, or

Freq Ref In, Sense

, or

Restore Input/Output Defaults.

The setting of

External Ref Coupling

is persistent across power-cycling and is not reset with a Preset. It is reset to the default state (

Normal

) when

Restore Input/Output Defaults

is invoked, which will also restart normal data acquisition.

The detection of invalid external reference is under interrupt processing. If the external reference becomes invalid then returns to valid in too short a time, no error condition will be detected or reported and therefore the acquisition will not be stopped.

Remote Command

Initial S/W Revision

Key Path

Mode

Preset

State Saved

Readback

[:SENSe]:ROSCillator:COUPling NORMal|NACQuisition

[:SENSe]:ROSCillator:COUPling?

A.02.00

Input/Output, Freq Ref In

All

This setting is persistent: it survives power-cycling or a Preset and is reset with Restore Input/Output defaults.

Not saved in State

Normal|Stop Acq

Output Config

Accesses keys that configure various output settings, like the frequency reference output, trigger output and analog output.

Key Path

Input/Output

89

Common Measurement Functions 1

Initial S/W Revision Prior to A.02.00

Trig Out (1 and 2)

Select the type of output signal that will be output from the rear panel Trig 1 Out or Trig 2 Out connectors.

Remote Command:

Example:

Dependencies:

Preset:

State Saved:

Key Path:

Initial S/W Revision:

:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut

HSWP|MEASuring|MAIN|GATE|GTRigger|OEVen|SSWeep|SSETtled

|S1Marker|S2Marker|S3Marker|S4Marker|OFF

:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut?

TRIG:OUTP HSWP

TRIG2:OUTP GATE

The second Trigger output (Trig 2 Out) does not appear in all models; in models that do not support it, the Trig 2 Out key is blanked, and sending the

SCPI command for this output generates an error, “Hardware missing; Not available for this model number” In models that do not support the Trigger 2 output, this error is returned if trying to set Trig 2 Out and a query of Trig 2

Out returns OFF.

Trigger 1: Sweeping (HSWP)

Trigger 2: Gate

This is unaffected by a Preset but is preset to the above values on a "Restore

Input/Output Defaults" or "Restore System Defaults->All"

Saved in instrument state

Input/Output, Output Config

Prior to A.02.00

Polarity

Sets the output to the Trig 1 Out or Trig 2 Out connector to trigger on either the positive or negative polarity.

Remote Command:

Example:

Preset:

:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut:POLarity

POSitive|NEGative

:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut:POLarity?

TRIG1:OUTP:POL POS

This is unaffected by a Preset but is set to POSitive on a "Restore Input/Output

Defaults" or "Restore System Defaults->All"

State Saved:

Key Path:

Initial S/W Revision:

Saved in instrument state

Input/Output, Output Config, Trig 1/2 Output

Prior to A.02.00

90

Common Measurement Functions 1

Sweeping (HSWP)

Selects the Sweeping Trigger signal to be output to the Trig 1 Out or Trig 2 Out connector when a measurement is made. This signal has historically been known as "HSWP" (High = Sweeping), and is 5 V TTL level with 50 ohm output impedance."

Example:

Key Path:

Readback:

Initial S/W Revision:

TRIG1:OUTP HSWP

Input/Output, Output Config, Trig 1/2 Output

Sweeping

Prior to A.02.00

Measuring

Selects the Measuring trigger signal to be output to the Trig 1 Out or Trig 2 Out connector. This signal is true while the Measuring status bit is true.

Example:

Key Path:

Readback:

Initial S/W Revision:

TRIG1:OUTP MEAS

Input/Output, Output Config, Trig 1/2 Output

Measuring

Prior to A.02.00

Main Trigger

Selects the current instrument trigger signal to be output to the Trig 1 Out or Trig 2 Out connector.

Example:

Key Path:

Readback:

Initial S/W Revision:

TRIG1:OUTP MAIN

Input/Output, Output Config, Trig 1/2 Output

Main Trigger

Prior to A.02.00

Gate Trigger

Selects the gate trigger signal to be output to the Trig 1 Out or Trig 2 Out connector. This is the source of the gate timing, not the actual gate signal.

Example:

Key Path:

Readback:

Initial S/W Revision:

TRIG1:OUTP GTR

Input/Output, Output Config, Trig 1/2 Output

Gate Trigger

Prior to A.02.00

Gate

Selects the gate signal to be output to the Trig 1 Out or Trig 2 Out connector. The gate signal has been delayed and its length determined by delay and length settings. When the polarity is positive, a high on the Trig 1 Out or Trig 2

91

Common Measurement Functions 1

Out represents the time the gate is configured to pass the signal.

Example:

Key Path:

Readback:

Initial S/W Revision:

TRIG1:OUTP GATE

Input/Output, Output Config, Trig 1/2 Output

Gate

Prior to A.02.00

Odd/Even Trace Point

Selects either the odd or even trace points as the signal to be output to the Trig 1 Out or Trig 2 Out connector when performing swept spectrum analysis. When the polarity is positive, this output goes high during the time the analyzer is sweeping past the first point (Point 0) and every other following trace point. The opposite is true if the polarity is negative.

Example:

Key Path:

Readback:

Initial S/W Revision:

TRIG1:OUTP OEV

Input/Output, Output Config, Trig 1/2 Output

Odd/Even

Prior to A.02.00

Off

Selects no signal to be output to the Trig 1 Out or Trig 2 Out connector.

Example:

Key Path:

Readback:

Initial S/W Revision:

TRIG1:OUTP OFF

Input/Output, Output Config, Trig 1/2 Output

Off

Prior to A.02.00

Analog Out

This menu lets you control which signal is fed to the “Analog Out” connector on the analyzer rear panel.

See also

“More Information” on page 93

.

Remote Command:

Example:

:OUTPut:ANALog OFF|SVIDeo|LOGVideo|LINVideo|DAUDio

:OUTPut:ANALog?

OUTP:ANAL SVIDeo ! causes the analog output type to be Screen Video

Preset:

Preset:

OFF

This is unaffected by Preset but is set to DAUDio on a "Restore Input/Output

Defaults" or "Restore System Defaults->All

92

Common Measurement Functions 1

Backwards Compatibility Notes: Prior to A.04.00, OFF was the default functionality except when in the Analog

Demod application or with Tune and Listen, in which case it was DAUDio, and there was no selection menu. So for backwards compatibility, Auto

(:OUTP:ANAL:AUTO ON) will duplicate the prior behavior.

The DNWB and SANalyzer parameters, which were legal in PSA but perform no function in the X-Series, are accepted without error.

State Saved:

Key Path:

Saved in Input/Output State

Readback line:

Initial S/W Revision:

Input/Output, Output Config

1-of-N selection [variable]

A.04.00

More Information

The table below gives the range for each output.

Analog Out

Off

Screen Video

Log Video

Linear Video

Demod Audio

Nominal Range exc.

(10% overrange) Scale Factor

0 V

0 to 1 V open circuit

0 to 1 V terminated

0 to 1 V terminated

10%/division

1/(192.66 dB/V)

100%/V

Notes

8566 compatible dB referenced to mixer level, 1 V out for –10 dBm at the mixer.

Linear referenced to Ref Level, 1 V out for RF envelope at the Ref

Level.

(varies with analyzer setting)

Auto

Selects the Auto state for the Analog Output menu. In this state, the Analog Output will automatically be set to the most sensible setting for the current mode or measurement.

If you make a selection manually from the Analog Out menu, this selection will remain in force until you change it

(or re-select Auto), even if you go to a mode or measurement for which the selected output does not apply.

Remote Command:

Example:

Preset:

:OUTPut:ANALog:AUTO OFF|ON|0|1

:OUTPut:ANALog:AUTO?

OUTP:ANAL:AUTO ON

ON

93

Common Measurement Functions 1

State Saved:

Key Path:

Initial S/W Revision:

Off

Turns off the analog output.

Example:

Key Path:

Readback Text:

Initial S/W Revision:

Saved in Input/Output State

Input/Output, Output Config, Analog Out

A.04.00

OUTP:ANAL OFF ! causes the analog output to be off

Input/Output, Output Config, Analog Out

Off

A.04.00

Screen Video

Selects the analog output to be the screen video signal. In this mode, the pre-detector data is output to the Analog

Out connector. The output looks very much like the trace displayed on the analyzer’s screen, and depends on the

Log/Lin display Scale, Reference Level, and dB per division, but is not influenced by the selected detector or any digital flatness corrections or trace post-processing (like Trace Averaging).

Note that this mode is similar to the Analog Output of the HP 8566 family and the Video Out (opt 124) capability of the Agilent PSA analyzer (E444x), although there are differences in the behavior.

See “Backwards Compatibility:” on page 95

.

Example:

Dependencies:

Couplings:

Key Path:

Readback Text:

Initial S/W Revision:

OUTP:ANAL SVID

Because the Screen Video output uses one of the two IF processing channels, only one detector is available while Screen Video is selected. All active traces will change to use the same detector as the selected trace when Screen Video is activated.

Screen Video output is not available while any EMI Detector is selected

(Quasi Peak, RMS Average or EMI Average), because these detectors use both IF processing channels. Consequently, if the user chooses an EMI

Detector, there will be no Screen Video output.

The output holds at its last value during an alignment and during a marker count and during retrace (after a sweep and before the next sweep starts).

This function depends on optional capability; the key will be blanked and the command will generate an “Option not available” error unless you have

Option YAV or YAS licensed in your instrument.

Screen Video output changes while in FFT Sweeps, so for measurements that use exclusively FFT Sweeps, or if the user manually chooses FFT Sweeps, the

Screen Video output will look different than it does in swept mode.

Input/Output, Output Config, Analog Out

Screen Video

A.04.00

94

Common Measurement Functions 1

Backwards Compatibility:

The Screen Video function is intended to be very similar to the 8566 Video Output and the PSA Option 124.

However, unlike the PSA, it is not always on; it must be switched on by the Screen Video key. Also, unlike the

PSA, there are certain dependencies (detailed above) – for example, the Quasi Peak Detector is unavailable when

Screen Video is on.

Futhermore, the PSA Option 124 hardware was unipolar and its large range was padded to be exactly right for use as a Screen Video output. In the X-Series, the hardware is bipolar and has a wider range to accommodate the other output choices. Therefore, the outputs won’t match up exactly and users may have to modify their setup when applying the X-Series in a PSA application.

Log Video

Selects the analog output to be the log of the video signal. In this mode, the pre-detector data is output to the

Analog Out connector with a Log scaling. The output does not depend on display settings like Reference Level or dB per division, and it is not influenced by the selected detector or any digital flatness corrections or trace post-processing (like Trace Averaging), but does change with input attenuation.

The output is designed so that full scale (1 V) corresponds to –10 dBm at the mixer. The full range (0–1 V) covers

192.66 dB ; thus, 0 V corresponds to –202.66 dBm at the mixer.

Example:

Dependencies:

Couplings:

Key Path:

Readback Text:

Initial S/W Revision:

OUTP:ANAL LOGV

Because the Log Video output uses one of the two IF processing channels, only one detector is available while Screen Video is selected. All active traces will change to use the same detector as the selected trace when Log Video is activated.

Log Video output is not available while any EMI Detector is selected (Quasi

Peak, RMS Average or EMI Average), because these detectors use both IF processing channels. Consequently, if the user chooses an EMI Detector, there will be no Log Video output.

The output holds at its last value during an alignment, during a marker count, and during retrace (after a sweep and before the next sweep starts).

This function depends on optional capability. The key will be blanked and the command will generate an “Option not available” error unless you have

Option YAV licensed in your instrument.

Log Video output changes while in FFT Sweeps, so for measurements that use exclusively FFT Sweeps, or if the user manually chooses FFT Sweeps, the

Log Video output will look different than it does in swept mode.

Input/Output, Output Config, Analog Out

Log Video

A.04.00

Linear Video

Selects the analog output to be the envelope signal on a linear (voltage) scale. In this mode, the pre-detector data is output to the Analog Out connector with a Linear scaling. The output is not influenced by the selected detector or any digital flatness corrections or trace post-processing (like Trace Averaging).

95

Common Measurement Functions 1

The scaling is set so that 1 V output occurs with an instantaneous video level equal to the reference level, and 0 V occurs at the bottom of the graticule. This scaling gives you the ability to control the gain without having another setup control for the key. But it requires you to control the look of the display (the reference level) in order to control the analog output.

This mode is ideal for looking at Amplitude Modulated signals, as the linear envelope effectively demodulates the signal.

Example:

Dependencies:

Couplings:

Key Path:

Readback Text:

Initial S/W Revision:

OUTP:ANAL LINV

Because the Linear Video output uses one of the two IF processing channels, only one detector is available while Linear Video is selected. All active traces will change to use the same detector as the selected trace when Log Video is activated.

Linear Video output is not available while any EMI Detector is selected

(Quasi Peak, RMS Average or EMI Average), because these detectors use both IF processing channels. Consequently, if the user chooses an EMI

Detector, there will be no Linear Video output.

The output holds at its last value during an alignment and during a marker count and during retrace (after a sweep and before the next sweep starts).

This function depends on optional capability; the key will be blanked and the command will generate an “Option not available” error unless you have

Option YAV licensed in your instrument.

Linear Video output changes while in FFT Sweeps, so for measurements that use exclusively FFT Sweeps, or if the user manually chooses FFT Sweeps, the

Linear Video output will look different than it does in swept mode.

Input/Output, Output Config, Analog Out

Linear Video

A.04.00

Demod Audio

Selects the analog output to be the demodulation of the video signal.

When Demod Audio is selected, the demodulated audio signal appears at this output whenever the Analog Demod application is demodulating a signal or when

Analog Demod Tune and Listen

is operating in the Swept SA measurement.

When Analog Out is in the Auto state, this output is auto-selected when in the Analog Demod mode or when

Analog Demod Tune and Listen

is operating in the Swept SA measurement.

If any other Analog Output is manually selected when in the Analog Demod mode or when

Analog Demod Tune and Listen

is operating in the Swept SA measurement, a warning message appears in the status message bar.

Example: OUTP:ANAL DAUD

96

Common Measurement Functions 1

Dependencies:

Key Path:

Readback Text:

Initial S/W Revision:

Modified at S/W Revision:

This key only appears if the Analog Demod application (N9063A) or Option

EMC is installed and licensed, otherwise the key will be blanked and the command will generate an “Option not available” error.

The output holds at its last value during an alignment and during a marker count. It is not held between sweeps, in order for Tune and Listen to work properly.

When Demod Audio is the selected Analog Output:

• all active traces are forced to use the same detector.

• CISPR detectors (QPD, EMI Avg, RMS Avg) are unavailable

Input/Output, Output Config, Analog Out

Demod Audio

Prior to A.02.00 (this was the default functionality, and there was no selection)

A.04.00

I/Q Cal Out

The Baseband I/Q "Cal Out" port can be turned on with either a 1 kHz or a 250 kHz square wave. This can be turned on independent of the input selection. A Preset will reset this to Off.

Remote Command

Example

Initial S/W Revision

Key Path

Couplings

Preset

State Saved

Range

Readback Text

:OUTPut:IQ:OUTPut IQ1|IQ250|OFF

:OUTPut:IQ:OUTPut?

OUTP:IQ:OUTP IQ1

Prior to A.02.00

Input/Output, Output Config

An I/Q Cable Calibration or an I/Q Probe Calibration will change the state of the Cal Out port as needed by the calibration routine. When the calibration is finished the I/Q Cal Out is restored to the pre-calibration state.

Off

Saved in instrument state.

1 kHz Square Wave|250 kHz Square Wave|Off

1 kHz|250 kHz|Off

1 kHz Square Wave

Turns on the 1 kHz square wave signal at the Cal Out port. This choice is only available with option BBA.

Key Path

Readback

Initial S/W Revision

Input/Output, Output Config, I/Q Cal Out

I/Q 1kHz

Prior to A.02.00

97

Common Measurement Functions 1

250 kHz Square Wave

Turns on the 250 kHz square wave signal at the Cal Out port. This choice is only available with option BBA.

Key Path

Readback

Initial S/W Revision

Input/Output, Output Config, I/Q Cal Out

I/Q 250kHz

Prior to A.02.00

Off

Turns off the signal at the Cal Out port. This choice is only available with option BBA.

Key Path

Readback

Initial S/W Revision

Input/Output, Output Config, I/Q Cal Out

Off

Prior to A.02.00

Digital Out

Opens a menu that allows you to select options for configuring the digital output(s) of the analyzer.

Initial S/W Revision Prior to A.02.00

Key Path

Input/Output, Output Config

Modified at S/W Revision A.04.00

Digital Bus (Narrowband)

This menu allows you to configure the LVDS connector located on the rear panel of the instrument. It is a unidirectional link of real time data at a 90 MSa/s rate. The ADC is sampling a 22.5 MHz IF.

The data that appears on this port is raw, uncorrected ADC samples, unless you have option RTL. With option

RTL, you get fully corrected I/Q data.

This connector will only be active when the Narrowband IF Path is currently in use.

Key Path

Initial S/W Revision

Input/Output, Output Config, Digital Out

A.04.00

Bus Out On/Off

When Bus Out is on, all acquisitions are streamed to the output port including acquisitions for internal purposes such as Alignment; internal processing and routing of acquisitions continues as usual and is unaffected by the state of Bus Out.

When Bus Out is off, no signal appears on the LVDS port.

Remote Command

:OUTPut:DBUS[1][:STATe] ON|OFF|1|0

:OUTPut:DBUS[1][:STATe]?

98

Common Measurement Functions 1

Example

Initial S/W Revision

Key Path

Scope

Preset

State Saved

Modified at S/W Revision

OUTP:DBUS ON

Prior to A.02.00

Input/Output, Output Config, Digital Out, Digital Bus

Mode Global

This is unaffected by a Preset but is set to Off on a "Restore Input/Output

Defaults" or "Restore System Defaults -> All"

Saved in Input/Output State

A.04.00

Aux IF Out

This menu controls the signals that appear on the SMA output on the rear panel labeled “AUX IF OUT”:

Remote Command:

:OUTPut:AUX SIF|AIF|LOGVideo|OFF

:OUTPut:AUX?

Preset: This is unaffected by a Preset but is set to OFF on a "Restore Input/Output

Defaults" or "Restore System Defaults->All”

Saved in Input/Output state State Saved:

Key Path:

Readback line:

Input/Output, Output Config

1-of-N selection [variable]

Backwards Compatibility Notes: In the PSA, the IF output has functionality equivalent to the "Second IF" function in the X-Series’ Aux IF Out menu. In the X-Series, it is necessary to switch the Aux IF Out to “Second IF” to get this functionaluity, whereas in

PSA it is always on, since there are no other choices. Hence a command to switch this function to “Second IF” will have to be added by customers migrating from PSA who use the IF Outputin PSA.

Initial S/W Revision: A.04.00

Second IF

In this mode the 2nd IF output is routed to the rear panel connector. The annotation on the key shows the current

2nd IF frequency in use in the analyzer.

The frequency of the 2nd IF depends on the current IF signal path as shown in the table below:

IF Path Selected

10 MHz

25 MHz

40 MHz

140 MHz

Frequency of “Second IF” Output

322.5 MHz

322.5 MHz

250 MHz

300 MHz

99

Common Measurement Functions 1

The signal quality, such as signal to noise ratio and phase noise, are excellent in this mode.

Example:

Key Path:

Readback Text:

Initial S/W Revision:

OUTP:AUX SIF causes the aux output type to be Second IF

Input/Output, Output Config, Aux IF Out

Second IF

A.04.00

Arbitrary IF

In this mode the 2nd IF output is mixed with a local oscillator and mixer to produce an arbitrary IF output between

10 MHz and 75 MHz with 500 kHz resolution. The phase noise in this mode will not be as good as in

Second IF

mode.

The IF output frequency is adjustable, through an active function which appears on the Arbitrary IF selection key, from 10 MHz to 75 MHz with 500 kHz resolution.

The bandwidth of this IF output varies with band and center frequency, but is about 40 MHz at the –3 dB width.

When the output is centered at lower freqeuncies in its range, signal frequencies at the bottom of the bandwidth will

“fold”. For example, with a 40 MHz bandwidth (20 MHz half-bandwidth), and a 15 MHz IF center, a signal –20

MHz relative to the spectrum analyzer center frequency will have a relative response of about –3 dB with a frequency 20 MHz below the 15 MHz IF center. This –5 MHz frequency will fold to become a +5 MHz signal at the IF output. Therefore, lower IF output frequencies are only useful with known band-limited signals.

Example:

Key Path:

Readback Text:

Initial S/W Revision:

OUTP:AUX AIF causes the aux output type to be the Arbitrary IF

Input/Output, Output Config, Aux IF Out

Arbitrary IF

A.04.00

Key Path

Scope

Remote Command

Example

Preset

State Saved

Min

Max

Default Unit

Input/Output, Output Config, Aux IF Out

Mode Global

:OUTPut:AUX:AIF <value>

:OUTPut:AUX:AIF?

:OUTP:AUX:AIF 50 MHZ

This is unaffected by a Preset but is set to 70 MHz on a "Restore Input/Output

Defaults" or "Restore System Defaults->All”

Saved in Input/Output State

10 MHz

75 MHz

Hz

100

Common Measurement Functions 1

Initial S/W Revision A.04.00

Fast Log Video

In this mode the 2nd IF output is passed through a log amp and the log envelope of the IF signal is sent to the rear panel. The open circuit output level varies by about 25 mV per dB, with a top-of-screen signal producing about 1.6

Volts. The output impedance is nominally 50 ohms.

This mode is intended to meet the same needs as Option E4440A-H7L Fast Rise Time Video Output on the Agilent

E4440A PSA Series, allowing you to characterize pulses with fast rise times using standard measurement suites on modern digital scopes.

Example:

Dependencies:

Key Path:

Readback Text:

Initial S/W Revision:

OUTP:AUX LOGVideo causes the aux output type to be Fast Log Video

The output is off during an alignment but not during a marker count, and is not blanked during retrace (after a sweep and before the next sweep starts).

Input/Output, Output Config, Aux IF Out

Fast Log Video

A.04.00

Off

In this mode nothing comes out of the “AUX IF OUT” connector on the rear panel. The connector appears as an open-circuit (that is, it is not terminated in any way).

Example:

Key Path:

Readback Text:

Initial S/W Revision:

OUTP:AUX OFF causes the aux output type to be off

Input/Output, Output Config, Aux IF Out

Off

A.04.00

I/Q Guided Calibration

Calibrating the Baseband I/Q ports requires several steps and manual connections. The Guided

Calibration will interactively step a user through the required steps, displaying diagrams to help with the connections. The steps will vary depending on the setup.

In the Guided Calibration windows, the date and time of the last calibration are displayed. If any of the items listed are displayed in yellow, this indicates that the calibration for that item is inconsistent with the latest calibration, and you should complete the entire calibration process before you exit the calibration.

I/Q Isolation Calibration

The I/Q Isolation Calibration must be run before calibrating any port with either the I/Q Cable

Calibration or I/Q Probe Calibration. This calibration is performed with nothing connected to any of the front panel I/Q ports. This is the first step in both the I/Q Cable Calibration and the I/Q Probe

101

Common Measurement Functions 1

Calibration.

Next

Perform the I/Q Isolation calibration.

Remote Command

Example

Initial S/W Revision

Key Path

Notes

Notes

:CALibration:IQ:ISOLation

CAL:IQ:ISOL

Prior to A.02.00

Input/Output, I/Q, I/Q Cable Calibration

All front panel I/Q ports must not be connected to anything.

All cables and probes should be disconnected from the I/Q ports before issuing the SCPI command.

No.

State Saved

Exit

Exits the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired calibration data.

Key Path

Notes

Initial S/W Revision

Input/Output, I/Q, I/Q Cable Calibration

Using the Exit button will not restore the calibration data to the state prior to entering the guided calibration. Once a port is calibrated the data is stored immediately and the only way to change it is to redo the calibration step.

When the calibration may be left in an inconsistent state, a confirmation

dialog will be displayed (see “Exit Confirmation” on page 113 ).

Prior to A.02.00

I/Q Isolation Calibration Time (Remote Command Only)

Returns the last date and time that the I/Q Isolation Calibration was performed. This is a remote query command only.

Remote Command:

Example:

Notes:

Initial S/W Revision:

:CALibration:IQ:ISOLation:TIME?

:CAL:IQ:ISOL:TIME?

This returns 6 integer values: year, month, day, hour, minute, second. When no calibration has been performed, all values will be 0.

A.02.00

I/Q Cable Calibrate…

The I/Q cable calibration creates correction data for each of the front panel I/Q ports. This calibration data is used whenever no probe specific calibration data is available. It is important that all ports are calibrated using the same short BNC cable so that the data is comparable from port to port.

102

Common Measurement Functions 1

The guided calibration (front panel only) will show connection diagrams and guide the user through the isolation calibration and calibrating each port. The calibration data for each port is stored separately, so as soon as a port is calibrated that data is saved and will be used. If a user presses "Exit" to exit the calibration process, the data for the ports already completed will still be used. It is recommended that a calibration be completed once started, or if exited, that it be properly done before the next use of the I/Q ports. The "Next" button will perform the calibration for the current port and then proceed to the next step in the calibration procedure. The "Back" button will return to the prior port in the procedure. Both softkeys and dialog buttons are supplied for ease of use. The dialog buttons are for mouse use and the softkeys for front panel use.

The calibration can also be done via SCPI, but no connection diagrams will be shown. The user will have to make the correct connections before issuing each port calibration command. Again, it is recommended that all ports be calibrated at the same time.

The instrument state remains as it was prior to entering the calibration procedure except while a port is actually being calibrated. Once a port is calibrated it returns to the prior state. A port calibration is in process only from the time the "Next" button is pressed until the next screen is shown. For SCPI, this corresponds to the time from issuing the CAL:IQ:FLAT:I|IB|Q|QB command until the operation is complete.

For example, if the prior instrument state is Cal Out = Off, Input = I+jQ, and Differential = Off, then up until the time the "Next" button is pressed the I Input and Q Input LEDs are on and the Cal Out, I-bar

Input and Q-bar Input LEDs are off. Once the "Next" button is pressed for the I port calibration, only the

Cal Out and I Input LEDs will be on and the others will be off. When the screen progresses to the next step ("Next" button again enabled), the prior state is restored and only the I Input and Q Input LEDs are on (Cal Out is off again).

The last calibration date and time for each port will be displayed. Any calibrations that are more than a day older than the most recent calibration will be displayed with the color amber.

Key Path

Initial S/W Revision

Input/Output, I/Q

Prior to A.02.00

I Port

The I port calibration is performed with the front panel's I port connected via a short BNC cable to the Cal Out port.

The guided calibration will show a diagram of the required connections.

Back

Return to the prior step in the calibration procedure.

Key Path

Initial S/W Revision

Next

Perform the I port calibration.

Remote Command

Input/Output, I/Q, Q Setup, Q Probe, Calibrate

Prior to A.02.00

:CALibration:IQ:FLATness:I

103

Common Measurement Functions 1

Example

Initial S/W Revision

Key Path

Notes

Notes

CAL:IQ:FLAT:I

Prior to A.02.00

Input/Output, I/Q, I/Q Cable Calibrate…

The recommended procedure is to use the same BNC cable to calibrate all I/Q ports. All I/Q ports should be calibrated sequentially during the procedure.

The calibration data is saved as soon as the port is calibrated and will survive power cycles. It is not reset by any preset or restore data commands.

The I port must be connected to the Cal Out port before issuing the SCPI command.

No.

State Saved

Exit

Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired calibration data.

Key Path

Notes

Initial S/W Revision

Input/Output, I/Q, I/Q Cable Calibrate…

Using the Exit button will not restore the calibration data to the state prior to entering the guided calibration. Once a port is calibrated the data is stored immediately and the only way to change it is to redo the calibration step.

When the calibration may be left in an inconsistent state, a confirmation

dialog will be displayed (see “Exit Confirmation” on page 113 ).

Prior to A.02.00

I-bar Port

The I-bar port calibration is performed with the front panel's I-bar port connected via a short BNC cable to the Cal

Out port. The guided calibration will show a diagram of the required connections.

Back

Return to the prior step in the calibration procedure.

Key Path

Notes

Initial S/W Revision

Next

Perform the I-bar port calibration.

Remote Command

Input/Output, I/Q, I/Q Cable Calibration

Using the Back button will not restore the calibration data to a prior state.

Once a port is calibrated the data is stored immediately and the only way to change it is to redo the calibration step. The Back button allows the user to go back to a prior step to redo that calibration step.

Prior to A.02.00

:CALibration:IQ:FLATness:IBAR

104

Common Measurement Functions 1

Example

Initial S/W Revision

Key Path

Notes

Notes

CAL:IQ:FLAT:IBAR

Prior to A.02.00

Input/Output, I/Q, I/Q Cable Calibrate…

The recommended procedure is to use the same BNC cable to calibrate all I/Q ports. All I/Q ports should be calibrated sequentially during the procedure.

The calibration data is saved as soon as the port is calibrated and will survive power cycles. It is not reset by any preset or restore data commands.

The I-bar port must be connected to the Cal Out port before issuing the SCPI command.

No State Saved

Exit

Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired calibration data.

Key Path

Notes

Initial S/W Revision

Input/Output, I/Q, I/Q Cable Calibrate…

Using the Exit button will not restore the calibration data to the state prior to entering the guided calibration. Once a port is calibrated the data is stored immediately and the only way to change it is to redo the calibration step.

When the calibration may be left in an inconsistent state, a confirmation dialog will be displayed (see

“Exit Confirmation” on page 113

).

Prior to A.02.00

Q Port

The Q port calibration is performed with the front panel's Q port connected via a short BNC cable to the Cal Out port. The guided calibration will show a diagram of the required connections.

Back

Return to the prior step in the calibration procedure.

Key Path

Notes

Input/Output, I/Q, I/Q Cable Calibrate…

Using the Back button will not restore the calibration data to a prior state.

Once a port is calibrated the data is stored immediately and the only way to change it is to redo the calibration step. The Back button allows the user to go back to a prior step to redo that calibration step.

Prior to A.02.00

Initial S/W Revision

Next

Perform the Q port calibration.

Remote Command

:CALibration:IQ:FLATness:Q

105

Common Measurement Functions 1

Example

Initial S/W Revision

Key Path

Notes

Notes

CAL:IQ:FLAT:Q

Prior to A.02.00

Input/Output, I/Q, I/Q Cable Calibrate…

The recommended procedure is to use the same BNC cable to calibrate all I/Q ports. All I/Q ports should be calibrated sequentially during the procedure.

The calibration data is saved as soon as the port is calibrated and will survive power cycles. It is not reset by any preset or restore data commands.

The Q port must be connected to the Cal Out port before issuing the SCPI command.

No State Saved

Exit

Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired calibration data.

Key Path

Notes

Initial S/W Revision

Input/Output, I/Q, I/Q Cable Calibrate…

Using the Exit button will not restore the calibration data to the state prior to entering the guided calibration. Once a port is calibrated the data is stored immediately and the only way to change it is to redo the calibration step.

When the calibration may be left in an inconsistent state, a confirmation

dialog will be displayed (see “Exit Confirmation” on page 113 ).

Prior to A.02.00

Q-bar Port

The Q-bar port calibration is performed with the front panel's Q-bar port connected via a short BNC cable to the

Cal Out port. The guided calibration will show a diagram of the required connections.

Back

Return to the prior step in the calibration procedure.

Key Path

Notes

Initial S/W Revision

Next

Perform the Q-bar port calibration.

Remote Command

Input/Output, I/Q, I/Q Cable Calibrate…

Using the Back button will not restore the calibration data to a prior state.

Once a port is calibrated the data is stored immediately and the only way to change it is to redo the calibration step. The Back button allows the user to go back to a prior step to redo that calibration step.

Prior to A.02.00

:CALibration:IQ:FLATness:QBAR

106

Common Measurement Functions 1

Example

Initial S/W Revision

Key Path

Notes

Notes

CAL:IQ:FLAT:QBAR

Prior to A.02.00

Input/Output, I/Q, I/Q Cable Calibrate…

The recommended procedure is to use the same BNC cable to calibrate all I/Q ports. All I/Q ports should be calibrated sequentially during the procedure.

The calibration data is saved as soon as the port is calibrated and will survive power cycles. It is not reset by any preset or restore data commands.

The Q-bar port must be connected to the Cal Out port before issuing the SCPI command.

No State Saved

Exit

Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired calibration data.

Key Path

Notes

Initial S/W Revision

Input/Output, I/Q, I/Q Cable Calibrate…

Using the Exit button will not restore the calibration data to the state prior to entering the guided calibration. Once a port is calibrated the data is stored immediately and the only way to change it is to redo the calibration step.

When the calibration may be left in an inconsistent state, a confirmation dialog will be displayed (see

“Exit Confirmation” on page 113

).

Prior to A.02.00

I/Q Cable Calibration Time (Remote Command Only)

Returns the last date and time that the I/Q Cable Calibration was performed for a specific port. This is a remote query command only.

Remote Command:

Example:

Notes:

:CALibration:IQ:FLATness:I|IBAR|Q|QBAR:TIME?

:CAL:IQ:FLAT:I:TIME?

This returns 6 integer values: year, month, day, hour, minute, second. When no calibration has been performed, all values will be 0.

A.02.00

Initial S/W Revision:

I/Q Probe Calibration

The I/Q probe calibration creates correction data for one of the front panel I/Q channels. When the probe has EEPROM identification, the data is unique to that specific probe. When the probe does not have

EEPROM identification, the data will be used for all probes of the same type. The data is also unique to the channel, so calibration data for the I channel will not be used for the Q channel and vice versa.

The guided calibration (front panel only) will show connection diagrams and guide the user through the

I/Q Isolation Calibration and through calibrating each port. The calibration data for each port is stored separately, so as soon as a port is calibrated that data is saved and will be used. If a user presses "Exit" to

107

Common Measurement Functions 1

exit the calibration process, the data for the port already completed will still be used. It is recommended that a calibration be completed once started, or if exited, that it be properly done before the next use of the probe. The "Next" button will perform the calibration for the current port and then proceed to the next step in the calibration procedure. The "Back" button will return to the prior port in the procedure.

Both softkeys and dialog buttons are supplied for ease of use. The dialog buttons are for mouse use and the softkeys for front panel use.

The calibration can also be done via SCPI, but no connection diagrams will be shown. The user will have to make the correct connections before issuing each port calibration command. Again, it is recommended that all ports be calibrated at the same time.

For Active probes or when Differential is Off, only the main port is calibrated, otherwise both the main and complementary ports are calibrated.

The instrument state remains as it was prior to entering the calibration procedure except while a port is actually being calibrated. Once a port is calibrated it returns to the prior state. A port calibration is in process only from the time the "Next" button is pressed until the next screen is shown. For SCPI, this corresponds to the time from issuing the CAL:IQ:PROB:I|IB|Q|QB command until the operation is complete.

For example, if the prior instrument state is Cal Out = Off, Input = I+jQ, and Differential = Off, then up until the time the "Next" button is pressed the I Input and Q Input LEDs are on and the Cal Out, I-bar

Input and Q-bar Input LEDs are off. Once the "Next" button is pressed for the I port calibration, only the

Cal Out and I Input LEDs will be on and the others will be off. When the screen progresses to the next step ("Next" button again enabled), the prior state is restored and only the I Input and Q Input LEDs are on (Cal Out is off again).

The last calibration date and time for each relevant port will be displayed. For passive probes with

Differential On, any calibration that is more than a day older than the most recent calibration will be displayed with the color amber.

I Port

The I port calibration is performed with the probe body attached to the front panel's I port and the probe tip connected via an adapter to the Cal Out port. The guided calibration will show a diagram of the required connections.

Show Adapter

Show a connection diagram and instructions for the probe and adapter. See

“Show Adapter Screen” on page 113

.

Key Path

Notes

Input/Output, I/Q, I Setup, I Probe, Calibrate

Either a passive or an active probe adapter diagram will be shown, depending on the type of probe attached.

Prior to A.02.00

Initial S/W Revision

Back

Return to the prior step in the calibration procedure.

Key Path

Input/Output, I/Q, Q Setup, Q Probe, Calibrate

108

Common Measurement Functions 1

Initial S/W Revision

Next

Perform the I port calibration.

Prior to A.02.00

Remote Command

Example

Initial S/W Revision

Key Path

Notes

:CALibration:IQ:PROBe:I

CAL:IQ:PROB:I

Prior to A.02.00

Input/Output, I/Q, I Setup, I Probe, Calibrate

The I port must be connected to the Cal Out port before issuing the SCPI command.

The calibration data is saved as soon as the port is calibrated and will survive power cycles. It is not reset by any preset or restore data commands.

No State Saved

Exit

Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired calibration data.

Key Path

Notes

Initial S/W Revision

Input/Output, I/Q, I Setup, I Probe, Calibrate

Using the Exit button will not restore the calibration data to the state prior to entering the guided calibration. Once a port is calibrated the data is stored immediately and the only way to change it is to redo the calibration step.

When the calibration may be left in an inconsistent state, a confirmation dialog will be displayed (see

“Exit Confirmation” on page 113

).

Prior to A.02.00

I-bar Port

The I-bar port calibration is performed with the probe body attached to the front panel's I-bar port and the probe tip connected via an adapter to the Cal Out port. The I-bar probe calibration is only available for passive probes with

Differential On. The guided calibration will show a diagram of the required connections.

Show Adapter

Show a connection diagram and instructions for the probe and adapter. See “Show Adapter Screen” on page 113 .

Key Path

Notes

Initial S/W Revision

Input/Output, I/Q, I Setup, I Probe, Calibrate

Either a passive or an active probe adapter diagram will be shown, depending on the type of probe attached.

Prior to A.02.00

109

Common Measurement Functions 1

Back

Return to the prior step in the calibration procedure.

Key Path

Notes

Initial S/W Revision

Next

Perform the I-bar port calibration.

Input/Output, I/Q, I Setup, I Probe, Calibrate

Using the Back button will not restore the calibration data to a prior state.

Once a port is calibrated the data is stored immediately and the only way to change it is to redo the calibration step. The Back button allows the user to go back to a prior step to redo that calibration step.

Prior to A.02.00

Remote Command

Example

Initial S/W Revision

Key Path

Notes

:CALibration:IQ:PROBe:IBar

CAL:IQ:PROB:IB

Prior to A.02.00

Input/Output, I/Q, I Setup, I Probe, Calibrate

The I-bar port must be connected to the Cal Out port before issuing the SCPI command.

The calibration data is saved as soon as the port is calibrated and will survive power cycles. It is not reset by any preset or restore data commands.

No State Saved

Exit

Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired calibration data.

Key Path

Notes

Initial S/W Revision

Input/Output, I/Q, I Setup, I Probe, Calibrate

Using the Exit button will not restore the calibration data to the state prior to entering the guided calibration. Once a port is calibrated the data is stored immediately and the only way to change it is to redo the calibration step.

When the calibration may be left in an inconsistent state, a confirmation

dialog will be displayed (see “Exit Confirmation” on page 113 ).

Prior to A.02.00

Q Port

The Q port calibration is performed with the probe body attached to the front panel's Q port and the probe tip connected via an adapter to the Cal Out port. The guided calibration will show a diagram of the required connections.

110

Common Measurement Functions 1

Show Adapter

Show a connection diagram and instructions for the probe and adapter. See “Show Adapter Screen” on page 113 .

Key Path

Notes

Input/Output, I/Q, Q Setup, Q Probe, Calibrate

Either a passive or an active probe adapter diagram will be shown, depending on the type of probe attached.

Prior to A.02.00

Initial S/W Revision

Back

Return to the prior step in the calibration procedure.

Key Path

Initial S/W Revision

Next

Perform the Q port calibration.

Input/Output, I/Q, Q Setup, Q Probe, Calibrate

Prior to A.02.00

Remote Command

Example

Initial S/W Revision

Key Path

Notes

:CALibration:IQ:PROBe:Q

CAL:IQ:PROB:Q

Prior to A.02.00

Input/Output, I/Q, Q Setup, Q Probe, Calibrate

The Q port must be connected to the Cal Out port before issuing the SCPI command.

The calibration data is saved as soon as the port is calibrated and will survive power cycles. It is not reset by any preset or restore data commands.

No State Saved

Exit

Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired calibration data.

Key Path

Notes

Initial S/W Revision

Input/Output, I/Q, Q Setup, Q Probe, Calibrate

Using the Exit button will not restore the calibration data to the state prior to entering the guided calibration. Once a port is calibrated the data is stored immediately and the only way to change it is to redo the calibration step.

When the calibration may be left in an inconsistent state, a confirmation dialog will be displayed (see

“Exit Confirmation” on page 113

).

Prior to A.02.00

111

Common Measurement Functions 1

Q-bar Port

The Q-bar port calibration is performed with the probe body attached to the front panel's Q-bar port and the probe tip connected via an adapter to the Cal Out port. The Q-bar probe calibration is only available for passive probes with Differential On. The guided calibration will show a diagram of the required connections.

Show Adapter

Show a connection diagram and instructions for the probe and adapter. See

“Show Adapter Screen” on page 113

.

Key Path

Notes

Input/Output, I/Q, Q Setup, Q Probe, Calibrate

Either a passive or an active probe adapter diagram will be shown, depending on the type of probe attached.

Prior to A.02.00

Initial S/W Revision

Back

Return to the prior step in the calibration procedure.

Key Path

Notes

Initial S/W Revision

Next

Perform the Q-bar port calibration.

Input/Output, I/Q, Q Setup, Q Probe, Calibrate

Using the Back button will not restore the calibration data to a prior state.

Once a port is calibrated the data is stored immediately and the only way to change it is to redo the calibration step. The Back button allows the user to go back to a prior step to redo that calibration step.

Prior to A.02.00

Remote Command

Example

Initial S/W Revision

Key Path

Notes

:CALibration:IQ:PROBe:QBar

CAL:IQ:PROB:QB

Prior to A.02.00

Input/Output, I/Q, Q Setup, Q Probe, Calibrate

The Q-bar port must be connected to the Cal Out port before issuing the SCPI command.

The calibration data is saved as soon as the port is calibrated and will survive power cycles. It is not reset by any preset or restore data commands.

No State Saved

Exit

Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired calibration data.

Key Path

Input/Output, I/Q, Q Setup, Q Probe, Calibrate

112

Common Measurement Functions 1

Notes

Initial S/W Revision

Using the Exit button will not restore the calibration data to the state prior to entering the guided calibration. Once a port is calibrated the data is stored immediately and the only way to change it is to redo the calibration step.

When the calibration may be left in an inconsistent state, a confirmation dialog will be displayed (see

“Exit Confirmation” on page 113

).

Prior to A.02.00

Show Adapter Screen

When one of the Probe Calibration Show Adapter buttons is pressed, a diagram of the probe with its adapter will be shown. Depending on the type of probe attached, either the Passive Probe Adapter or the Active Probe Adapter diagram will be shown.

I/Q Probe Calibration Time (Remote Command Only)

Return the last date and time that the I/Q Probe Calibration was performed for a specific port. This is a remote query command only.

Remote Command:

Example:

Notes:

Initial S/W Revision:

:CALibration:IQ:PROBe:I|IBAR|Q|QBAR:TIME?

:CAL:IQ:PROB:I:TIME?

This returns 6 integer values: year, month, day, hour, minute, second. When no calibration has been performed, all values will be 0. The value is specific to both the port and probe, so the value will change as probes are connected or disconnected.

A.02.00

Exit Confirmation

When Exit is pressed during one of the calibration routines, the calibration may be in an inconsistent state with some of the ports having newly measured calibration data and others with old data. If this is the case, a dialog box will appear to confirm that the user really wants to exit. A "Yes" answer will exit the calibration procedure, leaving potentially inconsistent calibration data in place. A "No" answer will return to the calibration procedure.

Marker

See “Marker Control Mode” on page 114 .

See “Setting the Marker X Axis Value” on page 114 .

See “Setting the Marker X Position in Trace Points” on page 115

.

See “Setting the Marker Y Axis Value” on page 116 .

The Marker key accesses the Marker menu. A marker can be placed on a trace to allow the value of the trace at the marker point to be determined precisely. The functions in this menu include a 1-of-N selection of the control mode Normal, Delta, Fixed, or Off for the selected marker. If the selected marker is Off, pressing Marker sets it to Normal and places it at the center of the screen on the trace determined by the Marker Trace rules.

Markers may also be used in pairs to read the difference (or delta) between two data points. They can be

113

Common Measurement Functions 1

used in Marker Functions to do advanced data processing, or to specify operating points in functions like

Signal Track and N dB Points.

The SCPI command in the table below selects the marker and sets the marker control mode as described under

Normal

,

Delta

,

Fixed

and

Off

, below. All interactions and dependencies detailed under the key description are enforced when the remote command is sent.

Remote Command:

Preset:

State Saved:

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MODE

POSition|DELTa|FIXed|OFF

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MODE?

OFF (all markers)

The marker control mode is saved in instrument state

Backwards Compatibility SCPI: :CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MODE SPAN|BAND

These parameters are aliased to POSition if sent.

A query does not reflect them.

Initial S/W Revision: Prior to A.02.00

Marker Control Mode

There are four control modes for markers:

Normal (POSition) - A marker that can be moved to any point on the X Axis by specifying its X Axis

value, and who's absolute Y Axis value (see section 8.1.4.7) is then the value of the trace point at that X

Axis value.

Delta (DELTa) - A marker that can be moved to any point on the X Axis by specifying its X Axis offset

from a reference marker, and whose absolute Y Axis value (see section 8.1.4.7) is then the value of the trace point at that X Axis value.

Fixed (FIXed) - A marker whose X Axis and Y Axis values may be directly or indirectly specified by

you, but whose Y Axis value remains fixed, once specified, and does not follow the trace. Fixed markers are useful as reference markers for Delta markers, as operands in a Peak Search operation, and as arbitrary reference points settable by you. These markers are represented on the display by an "X" rather then a diamond.

Off (OFF) - A marker which is not in use.

In the Swept SA measurement, the Preset control mode is Off for all markers.

Setting the Marker X Axis Value

The command below sets the marker X Axis value in the current marker X Axis Scale unit. In each case the marker that is addressed becomes the selected marker. It has no effect (other than to cause the marker to become selected) if the control mode is

Off, but

it is the SCPI equivalent of entering an X value if the control mode is

Normal, Delta,

or

Fixed

.

Remote Command:

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X <freq>

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X?

114

NOTE

Common Measurement Functions 1

Notes:

Preset:

Min:

Max:

If no suffix is sent it will use the fundamental units for the current marker X

Axis Scale. If a suffix is sent that does not match the current marker X Axis

Scale unit, an invalid suffix error will be generated.

If the specified marker is Fixed and a Marker Function is on, a message is generated. If the key is pressed, an advisory message is generated. If the equivalent SCPI command is sent, this same message is generated as part of a

“–221, Settings conflict” warning.

The query returns the marker’s absolute X Axis value if the control mode is

Normal

or

Fixed

. It returns the offset from the marker’s reference marker if the control mode is

Delta

. The query is returned in the fundamental units for the current marker X Axis scale: Hz for

Frequency

and

Inverse Time

, seconds for

Period

and

Time

. If the marker is

Off

the response is not a number.

After a preset, if X is queried with no value sent first, the center of screen value will be returned. This will depend on the frequency range of the instrument. 13.255 GHz is correct for the 26 GHz instruments only (Option

526).

∞ (minus infinity)

+

∞ (plus infinity). Unlike legacy analyzers, where the markers were forced to be on screen, X-Series marker values are not limited and do not clip

Default Unit: determined by X Axis Scale

Backwards Compatibility SCPI: :CALCulate:MARKer[1]|2|3|4:X:CENTer

Initial S/W Revision: Prior to A.02.00

Setting the Marker X Position in Trace Points

The command below sets the marker X position in trace points. It has no effect if the marker control mode is

Off. B

ut it is the SCPI equivalent of entering a value if the control mode is

Normal

or

Delta

or

Fixed

– except the setting is in trace points rather than X Axis Scale units.

The entered value in Trace Points is immediately translated into the current X Axis

Scale units for setting the value of the marker. The marker’s value in X Axis Scale

Units, NOT trace points, will be preserved if a change is made to the X Axis scale settings. Thus, if you use this command to place a marker on bucket 500, which happens at that time to correspond to 13 GHz, and then you change the Start

Frequency so that bucket 500 is no longer 13 GHz, the marker will stay at 13 GHz,

NOT at bucket 500! This is important to realize as it differs from the behavior of past Agilent analyzers.

115

Common Measurement Functions 1

Remote Command:

Notes:

Preset:

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X:POSitio n <real>

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X:POSitio n?

If the specified marker is Fixed and a Marker Function is on, a message is generated. If the key is pressed, an advisory message is generated. If the equivalent SCPI command is sent, this same message is generated as part of a

“–221, Settings conflict” warning.

The query returns the marker’s absolute X Axis value in trace points if the control mode is

Normal

or

Fixed

. It returns the offset from the marker’s reference marker in trace points if the control mode is

Delta

. The value is returned as a real number, not an integer, corresponding to the translation from

X Axis Scale units to trace points

After a preset, if X is queried with no value sent first, the center of screen value will be returned. So if per default, the number of Trace points is 1001, the center value will be 500.

Min:

Max:

0

Number of trace points – 1

Default Unit: unitless

Backwards Compatibility SCPI: :CALCulate:MARKer[1]|2|3|4:X:POSition:CENTer

Backwards Compatibility SCPI: The legacy command,

:CALCulate:MARKer[n]:X:POSition:CENTer <param> was used to control the centerpoint between the Delta and Reference marker in trace points (buckets) in Span Pair mode. In the new system, this is equivalent to simply setting the marker position in trace points. So this command is aliased to the command

:CALCulate:MARKer[n]:X:POSition <param>

NOTE

The UP/DOWN parameters will increment/decrement by one bucket. This will require a conversion to buckets and back.

Initial S/W Revision:

Remote Command:

Prior to A.02.00

Setting the Marker Y Axis Value

The command below selects the marker and sets the marker Y Axis value; the default unit is the current

Y Axis unit. It has no effect (other than selecting the marker) unless the marker control mode is

Fixed.

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Y <real>

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Y?

116

Common Measurement Functions 1

Example:

Preset:

Min:

Max:

Initial S/W Revision:

CALC:MARK2:MODE POS turns on marker 2 as a normal marker.

CALC:MARK2:X 20 GHZ moves marker 2 to 20 GHz if X Axis Scale is

Frequency. If X Axis Scale is Time, the –131 invalid suffix error is generated.

Trace value at center of screen. There is no way to predict what this will be after a preset.

∞ (minus infinity)

+

∞ (plus infinity)

Prior to A.02.00

Select Marker

Specifies the selected marker. The term “selected marker” is used throughout this document to specify which marker will be affected by the functions.

Key Path

Notes

Preset

State Saved

Initial S/W Revision

Marker

The selected marker is remembered even when not in the Marker menu and is used if a Search is done or a Band Function is turned on or for Signal Track or

Continuous Peak.

Marker 1

The number of the selected marker is saved in instrument state.

Prior to A.02.00

Normal

Sets the control mode for the selected marker to

Normal

and turns on the active function for setting its value. If the selected marker was

Off

, it is placed at the center of the screen on the trace specified by the marker’s Trace attribute.

A Normal mode

(POSition type) marker can be moved to any point on the X Axis by specifying its X

Axis value. Its absolute Y Axis value is then the value of the trace point at that X Axis value.

Key Path

Example

Notes

Couplings

State Saved

Initial S/W Revision

Marker

:CALC:MARK:MODE POS sets Marker 1 to Normal.

See the description under the “ Marker” key, above.

• The marker addressed by this command becomes the selected marker on the front panel.

The marker control mode (Normal, Delta, Fixed, Off) and X Axis value are saved in instrument state.

Prior to A.02.00

Delta

Sets the control mode for the selected marker to Delta and turns on the active function for setting its delta

117

Common Measurement Functions 1

value. If the selected marker was

Off

, it is placed at the center of the screen on the trace specified by the marker’s Trace attribute.

In Delta mode the marker result shows the relative result between the selected (Delta) marker and its reference marker. A delta marker can be moved to any point on the X Axis by specifying its X Axis offset from a reference marker. Its absolute Y Axis value is then the value of the trace point at that X

Axis value.

Key Path

Example

Notes

State Saved

Initial S/W Revision

Marker

:CALC:MARK:MODE DELT sets marker 1 to Delta.

See the description under the “ Marker” key, above.

The marker control mode (Normal, Delta, Fixed, Off) and X Axis value are saved in instrument state

Prior to A.02.00

Fixed

See

“Fixed Marker X Axis Value” on page 119 .

See

“Fixed Marker Y Axis Value” on page 119 .

Sets the control mode for the selected marker to Fixed. A fixed marker is fixed in the sense that it stays where you place it. It can be directly moved in both X and Y. It can be moved with a Peak Search. It can also be indirectly moved by re-zeroing the delta if it is a relative marker. If it is moved, it again becomes fixed at the X Axis point it moved to and it has a Y-axis result that it took on when it moved there. If a

Normal or Delta marker is changed to Fixed it becomes fixed at the X Axis point it was at, and with the

Y-axis result it had when it was set to Fixed.

In Fixed mode the marker result shows:

• If no Marker Function is on, the absolute X Axis and Y axis value of the marker

• If a Marker Function is on, the X Axis value and the Y-axis function result the marker had when it became fixed.

118

Common Measurement Functions 1

Fixed Marker X Axis Value

Key Path

Example

Notes

Dependencies

State Saved

Initial S/W Revision

Fixed Marker Y Axis Value

Marker, Fixed

:CALC:MARK:MODE FIX sets Marker 1 to Fixed.

See the description under the “ Marker” key, above.

You cannot directly set the X or Y value of a Fixed marker which has a marker function turned on. If an attempt is made to actually adjust it while a Marker

Function is on, a message is generated. If the key is pressed, an advisory message is generated. If the equivalent SCPI command is sent, this same message is generated as part of a “–221, Settings conflict” warning. you cannot directly set the Y value of a Fixed marker while Normalize is turned on. If an attempt is made to do so while Normalize is on, a message is generated. If the key is pressed, an advisory message is generated. If the equivalent SCPI command is sent, this same message is generated as part of a

“–221, Settings conflict” warning.

The marker control mode (Normal, Delta, Fixed, Off) and X and Y Axis values are saved in instrument state

Prior to A.02.00

Key Path

Example

Notes

Dependencies

State Saved

Default Unit

Initial S/W Revision

Marker, Fixed

:CALC:MARK:MODE FIX sets Marker 1 to Fixed.

See the description under the

Marker

key, above.

you cannot directly set the X or Y value of a Fixed marker which has a marker function turned on. If an attempt is made to actually adjust it while a Marker

Function is on, , a message is generated. If the key is pressed, an advisory message is generated. If the equivalent SCPI command is sent, this same message is generated as part of a “–221, Settings conflict” warning.

The marker control mode (Normal, Delta, Fixed, Off) and X and Y Axis values are saved in instrument state depends on the current selected Y axis unit

Prior to A.02.00

Off

Turns off the selected marker and its marker function setting, if any. However, Off does not affect which marker is selected.

Key Path

Example

Notes

Marker

:CALC:MARK:MODE OFF sets Marker 1 to Off.

See the description under the “ Marker” key, above.

119

Common Measurement Functions 1

State Saved The marker control mode (Normal, Delta, Fixed, Off) is saved in instrument state

Prior to A.02.00

Initial S/W Revision

Properties

Opens a menu used to set certain properties of the selected marker.

Key Path

Initial S/W Revision

Marker

Prior to A.02.00

Select Marker

Duplicate of the

Select Marker

key under

Marker

. Selecting a marker here causes the same marker to be selected under

Marker.

(That is, there is only one “selected marker”.)

Relative To

Selects the marker that the selected marker will be relative to (its reference marker).

Every marker has another marker to which it is relative. This marker is referred to as the “reference marker” for that marker. This attribute is set by the

Marker, Properties, Relative To

key. The marker must be a

Delta

marker to make this attribute relevant. If it is a

Delta

marker, the reference marker determines how the marker is controlled and how its value is displayed. A marker cannot be relative to itself.

Key Path

Remote Command

Example

Notes

Notes

Couplings

Marker, Properties

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:REFerence

<integer>

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:REFerence

?

CALC:MARK1:REF 2 sets the marker 1 reference marker to 2 and turns marker 1 on as a delta marker.

A marker cannot be relative to itself so that choice is grayed out. If the grayed out key is pressed, an advisory message is generated. If the equivalent SCPI command is sent, this same message is generated as part of a “–221, Settings conflict” warning. See error –221.2200 in Master Error Messages: X-Series document for exact error text.

This command causes the marker specified with the subopcode to become selected.

Range (for SCPI command): 1 to 12. If the range is exceeded the value is clipped.

The act of specifying the selected marker’s reference marker makes the selected marker a Delta marker. If the reference marker is off it is turned on in

Fixed

mode at the delta marker location.

120

Common Measurement Functions 1

Preset

State Saved

Min

Max

Status Bits/OPC dependencies

The preset default “Relative To” marker (reference marker) is the next higher numbered marker (current marker +1). For example, if marker 2 is selected, then it’s default reference marker is marker 3. The exception is marker 12, which has a default reference of marker 1.

Set to the defaults by using

Restore Mode Defaults

. This is not reset by

Marker Off

,

All Markers Off

, or

Preset

.

Saved in instrument state. Not affected by Marker Off and hence not affected by Preset or power cycle.

1

12 none

Default (selected when

Restore Mode Defaults

is pressed): next higher numbered marker or 1 if marker 12.

Prior to A.02.00

Initial S/W Revision

X Axis Scale (formerly Readout)

Accesses a menu that enables you to affect how the X Axis information for the selected marker is displayed in the marker area (top-right of display) and the active function area of the display, and how the marker is controlled. The available settings for the X Axis Scale are Frequency, Period, Time, and

Inverse Time.

See “More Information” on page 122

.

Key Path

Remote Command

Example

Notes

Notes

Preset

State Saved

Marker, Properties

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X:READout

FREQuency|TIME|ITIMe|PERiod

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X:READout

?

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X:READout

:AUTO ON|OFF|1|0

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X:READout

:AUTO?

CALC:MARK3:X:READ TIME sets the marker 3 X Axis Scale to Time.

This command causes the specified marker to become selected.

This command causes the specified marker to become selected.

AUTO

Marker Preset (selected when a marker is turned

Off

):

Auto

(see below). In most measurements the Auto settings results in Frequency being the preset readout.

Saved in instrument state

121

Common Measurement Functions 1

Initial S/W Revision Prior to A.02.00

More Information

The

X Axis Scale

of a marker is the scale of its X Axis value. This affects the units displayed in the

Marker Result block and used to specify the marker’s X Axis location. The X Axis Scale is specified using the

Marker, Properties, X Axis Scale

key.

All markers in swept spans have both a time and frequency value. Which of these is used for the result display, and for positioning the marker, depends on the

X Axis Scale

setting. The

X Axis Scale

setting can be

Frequency

or

Time

, as well as the reciprocal of either (

Period

or

Inverse Time

). There is also an

Auto

setting - when in

Auto

, a marker’s

X Axis Scale

changes whenever the domain of the trace, upon which it set, changes. All choices for

X Axis Scale

are allowed. Note that this behavior differs from the behavior in previous instruments: previously the instrument remembered a different

X Axis Scale

(formerly called

Readout

) for each domain, and the choices of

X Axis Scale

were restricted. These restrictions were based on the current domain of the instrument.

Auto

When in Auto, the X-Axis Scale is

Frequency

if the Marker Trace is a frequency domain trace,

Time

if the Marker

Trace is a time domain trace. When in Auto, if the marker changes traces, or the domain of the trace the marker is on changes, the auto result is re-evaluated. If the X Axis Scale is chosen manually, that Scale is used regardless of the domain of the trace.

Key Path

Example

Marker, Properties, X Axis Scale

CALC:MARK2:X:READ:AUTO ON sets the marker 2 X-axis scaling to automatically select the most appropriate units.

Prior to A.02.00

Initial S/W Revision

Frequency

Sets the marker X Axis scale to Frequency, displaying the absolute frequency of a normal marker or the frequency of the delta marker relative to the reference marker. Frequency is the auto setting for frequency domain traces.

If Frequency is selected for a time domain trace, all of the points in the trace will show the same value. Attempting to use the knob or step keys to adjust the X Axis value of the marker or entering an X Axis value from the numeric keypad or remotely will have no effect but will generate no error.

Key Path

Example

Notes

State Saved

Initial S/W Revision

Marker, Properties, X Axis Scale

CALC:MARK2:X:READ FREQ !sets the marker 2 X Axis scale to

Frequency.

1-of-N readback is

Frequency

The X Axis Scale setting is saved in instrument state.

Prior to A.02.00

Period

Sets the marker X Axis scale to Period, displaying the reciprocal of the frequency of the marker, or the reciprocal of the frequency separation of the two markers in a delta-marker mode. The units are those of time (sec, msec, etc). If

122

Common Measurement Functions 1

the markers are at the same frequency in a delta marker mode, the result will be the reciprocal of 0, which is infinitely large. The display will show “---” and a SCPI query will return infinity.

If Period is selected for a time domain trace, all of the points in the trace will show the same value. Attempting to use the knob or step keys to adjust the X Axis value of the marker or entering an X Axis value from the numeric keypad or remotely will have no effect but will generate no error.

Key Path

Example

Notes

State Saved

Initial S/W Revision

Marker, Properties, X Axis Scale

CALC:MARK2:X:READ PER sets the marker 2 X Axis scale to Period.

1-of-N readback is

Period

The X Axis Scale setting is saved in instrument state.

Prior to A.02.00

Time

Sets the marker X Axis scale to Time, displaying the time interval between a normal marker and the start of a sweep or the time of the delta marker relative to the reference marker. Time is the auto setting for time domain traces. In a delta-marker mode it is the (sweep) time interval between the two markers.

Key Path

Example

Notes

Couplings

Marker, Properties, X Axis Scale

CALC:MARK2:X:READ TIME sets the marker 2 X Axis Scale to Time..

1-of-N readback is

Time

Frequency domain traces taken in FFT mode have no valid time data.

Therefore when Time is selected for markers on such traces, the X Axis value is taken as the appropriate percentage of the displayed sweep time, which is a calculated estimate.

The X Axis Scale setting is saved in instrument state

Prior to A.02.00

State Saved

Initial S/W Revision

Inverse Time

Sets the marker X Axis scale to Inverse Time, displaying the reciprocal time. It is useful in a delta mode to show the reciprocal of (sweep) time between two markers. This function is only meaningful when on a time domain trace and in the

Delta

control mode. If the markers are at the same X Axis value, the time between them is 0, so the reciprocal of sweep time is infinitely large. The display will show “---” and a SCPI query will return infinity.

Key Path

Example

Notes

Couplings

State Saved

Marker, Properties, X Axis Scale

:CALC:MARK2:X:READ ITIM sets the marker 2 X Axis scale to Inverse

Time.

1-of-N readback is

Inverse Time

Frequency domain traces taken in FFT mode have no valid time data.

Therefore when Inverse Time is selected for markers on such traces, the X

Axis value is undefined, shows as “---” and returns not a number to a query.

The X Axis Scale setting is saved in instrument state

123

Common Measurement Functions 1

Initial S/W Revision Prior to A.02.00

Marker Trace

Selects the trace that you want your marker to be placed on. A marker is associated with one and only one trace. This trace is used to determine the placement, result, and X Axis Scale of the marker. All markers have an associated trace, even

Fixed

markers; it is from that trace that they determine their attributes and behaviors, and it is to that trace that they go when they become Normal or Delta markers.

See

“Auto Init On” on page 124

.

See

“Auto Init Rules Flowchart” on page 125 .

See

“Auto Init OFF” on page 125

.

Key Path

Remote Command

Example

Notes

Couplings

Preset

State Saved

Min

Max

Readback line

Marker, Properties

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:TRACe

1|2|3|4|5|6

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:TRACe?

CALC:MARK1:TRAC 2 places marker 1 on trace 2.

A marker may be placed on a blanked and/or inactive trace, even though the trace is not visible and/or updating.

An application may register a trace name to be displayed on the key instead of a trace number.

1

6

The state of Marker Trace is not affected by the Auto Couple key.

If a Marker Trace is chosen manually, Auto Init goes to Off for that marker.

Sending the remote command causes the addressed marker to become selected.

Presets on Preset or All Markers Off

The Marker Trace and state of

Auto Init

for each marker is saved in instrument state.

[TraceN, Auto Init] or [TraceN, Manual] where N is the trace number to which the marker is currently assigned.

Prior to A.02.00

Initial S/W Revision

Auto Init On

When

Auto Init

is true, the marker’s trace attribute is re-determined automatically by the analyzer whenever the marker turns on (Normal, Delta or Fixed) from an Off state. (The trace attribute is also determined for all markers that are on, whenever

Auto Init

is turned on).

When the marker moves between traces the marker’s X position in trace points is retained as it moves.

124

Common Measurement Functions 1

For moving between active traces this generally means the x-axis value of the marker will not change.

But for moving to or from an inactive trace, the x-axis value will take on that of the new trace at the bucket the marker was on the old trace (and is still on, on the new trace, since the bucket doesn’t change).

Note this is true even if the marker is off screen. Thus, a marker that is at the center of the screen on the old trace stays at the center of the screen on the new trace. A marker that is off screen one whole screen to the left on the old trace remains off screen one whole screen to the left on the new trace – even if this means it will be at negative time!

Marker Trace is set to 1, and Auto Init is set to On, on a Preset or All Markers Off.

Auto Init Rules Flowchart

The following flowchart depicts the Auto Init rules:

Any trace have both Update and

Display

on?

YES

NO

Any trace have

Display on?

NO

YES

Any trace have

Update on?

NO

YES

Use lowest numbered such trace

Use lowest numbered such trace

Use lowest numbered such trace

Use trace 1

This flowchart makes it clear that putting all lower-numbered traces in View is the simplest way to specify which trace you want the markers to go to when they turn on. For example, if you want all

Markers to go to trace 2 when they turn on, put trace 1 in View.

Auto Init OFF

This command associates the marker with the specified trace and turns Marker Trace, Auto Init OFF for that marker. If the marker is not

Off

it moves the marker from the trace it was on to the new trace. If the marker is

Off

it stays off but is now associated with the specified trace.

The query returns the number of the trace on which the marker is currently placed, even if that marker is

125

Common Measurement Functions 1

in Auto mode.

Remote Command:

Notes:

Couplings:

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:TRACe:AUT

O OFF|ON|0|1

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:TRACe:AUT

O?

Turning Marker Trace Auto Init off has no effect on the trace on which the marker is currently placed.

The response to the query will be 0 if OFF, 1 if ON.

The state of Auto Init is not affected by the Auto Couple key.

Auto Init is set to True on a Preset or All Markers Off.

If Auto Init is set to On for a marker and that marker is on, that marker’s

Marker Trace is immediately set according to the above flowchart.

Sending the remote command causes the addressed marker to become selected.

ON

Prior to A.02.00

Preset:

Initial S/W Revision:

Lines

When on, displays a vertical line of graticule height and a horizontal line of graticule width, intersecting at the indicator point of the marker (that is, the center of the X or the bottom tip of the diamond. The lines are blue in color.

If the marker is off screen the lines should be extended from the marker so that they go thru the screen area if possible. This is really useful for off screen Fixed markers as it lets you see their amplitude even though they are off the X Axis.

Key Path

Remote Command

Example

Couplings

Preset

State Saved

Initial S/W Revision

Marker, Properties

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:LINes[:ST

ATe] OFF|ON|0|1

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:LINes[:ST

ATe]?

:CALC:MARK2:LIN:ON turns Lines on for marker 2.

Sending the remote command causes the addressed marker to become selected.

OFF

Saved in State

Prior to A.02.00

Marker Table

When set to On the display is split into a measurement window and a marker data display window. For

126

Common Measurement Functions 1

each marker which is on, information is displayed in the data display window, which includes the marker number, control mode, trace number, X axis scale, X axis value, and the Y-axis result. Additional information is shown for markers which have marker functions turned on.

Turning the Marker Table on turns the Peak Table off and vice versa.

Key Path

Remote Command

Example

Preset

State Saved

Initial S/W Revision

Marker

:CALCulate:MARKer:TABLe[:STATe] OFF|ON|0|1

:CALCulate:MARKer:TABLe[:STATe]?

CALC:MARK:TABL ON turns on the marker table.

OFF

Whether the marker table is on is saved in instrument state

Prior to A.02.00

Marker Count

Accesses the marker count menu.

Key Path

Readback line

Initial S/W Revision

Marker

[On] if count on for the selected marker, [Off] if it is off.

Prior to A.02.00

Counter On/Off

Turns the marker frequency counter on and off. The selected marker is counted, and if the selected marker is a delta marker and its reference marker is not fixed, the reference marker is counted as well.

See “Understanding the Marker Counter” on page 129 .

See “Query Count Value” on page 128 .

Key Path

Remote Command

Example

Marker Fctn, Marker Count

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt[:S

TATe] OFF|ON|0|1

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt[:S

TATe]?

CALC:MARK2:FCO ON selects marker 2, turns it on, and turns on the counter

CALC:MARK2:FCO:X? returns the counted frequency.

127

Common Measurement Functions 1

Notes

Notes

Dependencies

Couplings

Preset

State Saved

Initial S/W Revision

Fixed markers are not counted, but a Fixed marker will have a count stored in it if it is selected or is the reference marker for the selected marker. The count already in the marker is stored when the marker becomes fixed and if there is none or the marker moves (for example, Pk Search) it is counted and stored after the next sweep.

If a Fixed marker has a count stored in it, that count will be displayed when the marker is selected, and used as the reference count when that marker is a reference marker.

If a Fixed marker has a count stored in it, that count will be deleted if the marker X is adjusted.

If a Fixed marker has a count stored in it, and a Search function is performed using the Fixed marker, while the counter is on, the count stored in the marker will be updated.

If a Fixed marker has a count stored in it, and is a reference marker, and the reference is moved to a valid trace point by re-zeroing the delta (by pressing

Delta again or sending the DELTa SCPI command), while the counter is on, the count stored in the marker will be updated.

This command causes the specified marker to become selected.

Marker Count is unavailable (grayed out and Off) if the Gate function is on.

If the selected marker is

Off

when the counter is turned on, the selected marker is set to Normal and placed at center of screen on the trace determined by the

Marker Trace rules.

If a marker which is OFF is selected while the counter is on, the counter remains on, but since the marker is off, the count is undefined. In this case the analyzer will return not a number to a SCPI count query.

The counter is turned OFF when the selected marker is turned OFF.

OFF

The state of the counter (on/off) is saved in instrument state. In the case of

Fixed markers, the count stored in the marker is saved in instrument state.

Prior to A.02.00

Query Count Value

Queries the frequency count. The query returns the absolute count unless the specified marker is in Delta mode, then it returns the relative count. If the marker is off, or the marker is on but the counter is off, the analyzer will return not a number to a SCPI count query. A marker with no stored count, or a non-

Fixed

marker on a stored trace, will also return not a number to a SCPI count query. Note this result may simply mean that the first sweep after the counter turned on has not yet completed.

Remote Command:

Notes:

Initial S/W Revision:

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt:X?

This query does NOT cause the specified marker to become selected.

Prior to A.02.00

128

Common Measurement Functions 1

Understanding the Marker Counter

See “Counting Off-screen Markers” on page 129

.

See “Delta Marker” on page 129

.

See “Fixed Markers” on page 129

.

See “More Information on "Counter"” on page 130 .

Using the internal counter we can count the frequency of a marker, but we cannot count while we are actually sweeping. So, once we are done with a sweep, we move to the selected marker frequency and count that frequency. Then, if the marker is a Delta marker, the count is also taken for its reference marker. The count is actually performed by moving the LO to the frequency (or frequencies in the case of a delta marker) we wish to count. The count is executed on a marker by marker basis and no further count is taken until after the next sweep (even if the marker moves before another sweep has completed).

The Marker Count is taken by tuning the instrument to the frequency of the marker and counting the IF, with the instrument not sweeping. The count is adjusted for display by adding or subtracting it (as appropriate) from the LO frequency, so that you see a count that represents the signal frequency. This is true even if External Mixing is on. Since all this happens between sweeps, you never see the instrument retuning to do the counts.

If you wish to see the entered frequency of a counted marker it will appear in the active function area when that marker is selected (for Fixed markers, you have to press the Marker, Fixed key to select Fixed markers and then press it a second time to view or adjust the x or y marker values).

Counting Off-screen Markers

If the selected marker is off the X-axis the instrument can still be tuned to the marker (unless it is outside the range of the instrument), so the count can still be displayed. This means you can see a count for an off-screen marker even though there may be no valid Y-value for the marker. If the marker frequency is outside the range of the instrument, the display will show three dashes in the count block (---), and not a number is returned to a SCPI count query.

Delta Marker

When a Delta Marker is selected while Marker Count is on:

If the reference marker is not a fixed marker, the display shows the difference between the count of the selected marker and the count of the reference marker

If the reference marker is a fixed marker and there is a count stored in the marker (because Marker Count was on when the marker became a fixed marker), the display shows the difference between the count at the marker and the count stored in the reference marker.

Marker Count works in zero span as well as in Swept SA. The instrument tunes to the frequency of the selected marker, which, for active zero span traces, is simply the center frequency of the analyzer.

Fixed Markers

Fixed markers have a count stored in them that is generally kept fixed and not updated. If a fixed marker is selected, or used as a reference, the signal at the marker frequency is not counted; rather the stored count is seen or used as the reference. The count is stored, if Count is on, when the marker becomes fixed

129

Common Measurement Functions 1

or when, while fixed, the marker is moved by re-zeroing the reference (if it is the reference marker) or via a peak search (since both of these, by definition, use valid trace data). The count stored in a Fixed marker is lost if the counter is turned off, if the marker is moved to an inactive trace, or if the marker is moved by adjusting its x-value.

More Information on "Counter"

When the counter is on, the count (or the delta count) for the selected marker is displayed.

The invalid data indicator (*) will turn on until the completion of the first count.

Marker Count frequency readings are corrected using the

Freq Offset

function (in some previous analyzers, they were not). Note however that Marker Delta readings are not corrected, as any offset would be applied to both.

In zero span on active traces the counter continues to function, counting any signal near the center frequency of the analyzer.

NOTE

No signal farther from the marker frequency than the Res BW will be seen by the counter.

The above command turns on or off the frequency counter. If the specified marker number in the command is not the selected marker, it becomes the selected marker. If the specified marker number is not on, FCOunt ON sets it to Normal and places it at center of screen on the trace determined by the

Marker Trace rules. Once the marker count is on, it is on for any selected marker, not just for the one used in the command. A 1 is returned to the state query only if marker count is on and the specified number is the selected marker. The invalid data indicator (*) will turn on until the completion of the first count but this does not keep a value from being returned.

Gate Time Auto/Man

Controls the length of time during which the frequency counter measures the signal frequency. Longer gate times allow for greater averaging of signals whose frequency is “noisy”, though the measurement takes longer. If the gate time is an integer multiple of the length of a power-line cycle (20 ms for 50 Hz power, 16.67 ms for 60 Hz power), the counter rejects incidental modulation at the power line rate. The shortest gate time that rejects both 50 and 60 Hz modulation is 100 ms, which is the value chosen in

Auto, or on Preset or when Auto Couple is pressed.

The start time of the Gate Time of the counter must be controlled by the same trigger parameters as controls the sweep. Thus, if the Trigger is not in Free Run, the counter gate must not start until after the trigger is received and delayed.

Key Path

Marker Function, Marker Count

130

Common Measurement Functions 1

Remote Command

Example

Notes

Notes

Preset

State Saved

Min

Max

Initial S/W Revision

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt:GA

Tetime <time>

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt:GA

Tetime?

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt:GA

Tetime:AUTO OFF|ON|0|1

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FCOunt:GA

Tetime:AUTO?

:CALC:MARK2:FCO:GAT 1e–2 sets the gate time for Marker 2 to 10^(–2) s

= 10 ms.

When Auto Couple is pressed, Gate Time is set to 100 ms.

This command causes the specified marker to become selected.

100 ms

ON

Saved in instrument state.

1 us

500 ms

Prior to A.02.00

Couple Markers

When this function is true, moving any marker causes an equal X Axis movement of every other marker which is not Fixed or Off. By “equal X Axis movement” we mean that we preserve the difference between each marker’s X Axis value (in the fundamental x-axis units of the trace that marker is on) and the X Axis value of the marker being moved (in the same fundamental x-axis units).

Note that Fixed markers do not couple. They stay where they were while all the other markers move. Of course, if a Fixed marker is being moved, all the non-fixed markers do move with it.

This may result in markers going off screen.

Key Path

Remote Command

Example

Preset

State Saved

Initial S/W Revision

Marker

:CALCulate:MARKer:COUPle[:STATe] OFF|ON|0|1

:CALCulate:MARKer:COUPle[:STATe]?

:CALC:MARK:COUP ON sets Couple Markers on.

Off, presets on Mode Preset and All Markers Off

Saved in State

Prior to A.02.00

131

Common Measurement Functions 1

All Markers Off

Turns off all markers. See Marker,

“Off ” on page 119

.

Key Path

Remote Command

Example

Couplings

Preset

Initial S/W Revision

Marker

:CALCulate:MARKer:AOFF

CALC:MARK:AOFF turns off all markers. sets the selected marker to 1.

n/a.

Prior to A.02.00

Marker Function

The Marker Function key opens up a menu of softkeys that allow you to control the Marker Functions of the instrument. Marker Functions perform post-processing operations on marker data. Band Functions are Marker Functions that allow you to define a band of frequencies around the marker. The band defines the region of data used for the numerical calculations. These marker functions also allow you to perform mathematical calculations on trace and marker data and report the results of these calculations in place of the normal marker result.

NOTE

Unlike regular markers, marker function markers are not placed directly on the trace. They are placed at a location which is relative to the result of the function calculation.

See

“More Information” on page 133 .

See

“Fixed marker functions” on page 133 .

See

“Interval Markers” on page 133 .

Remote Command:

Notes:

Dependencies:

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion

NOISe|BPOWer|BDENsity|OFF

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion?

Sending this command selects the subopcoded marker

The marker function result is queried in the same fashion as the Marker

Result, as outlined in the Marker section, with the CALC:MARK:Y? command.

Fixed markers: It is not possible to change the Band Function for a Fixed marker; so all of the Band Function keys are grayed out for a Fixed marker.

If a marker function was already on when the marker became Fixed then the selected Band Function is shown but cannot be changed. Therefore, you cannot directly set the X or Y value of a Fixed marker which has a marker function turned on. To turn off the function, turn off the marker.

132

Common Measurement Functions 1

Preset:

State Saved:

Initial S/W Revision:

OFF

The band function for each marker is saved in Instrument State

Prior to A.02.00

More Information

The units to be used for displaying Marker Function results in Delta mode vary depending on what is the reference marker and what it is referenced to.

Marker Functions are different from Measurements, which automatically perform complex sequences of setup, data acquisition, and display operations in order to measure specified signal characteristics.

Marker Functions are specified for each individual marker and may be turned on individually for each marker.

The

Marker Fctn

menu controls which marker functions are turned on and allows you to adjust setup parameters for each function. The Marker Functions are

Marker Noise

,

Band/Interval Power

, and

Band/Interval Density,

only one of which can be on for a given marker.

If the selected marker is off, pressing Marker Fctn sets it to Normal and places it at the center of the display on the trace determined by the Marker Trace rules. However, if the selected marker was

Off

,

Marker Function Off

had to be the selected function, and it remains so even after the marker is thus turned on, although you may then change it.

Fixed marker functions

In the case of a fixed marker, it is not possible to turn on or change a band function. This is because a

Fixed marker holds the value it had when it became fixed; the trace it was on may keep on changing, so the function value, which depends on trace data, could not be calculated on an ongoing basis.

It is possible to have a Marker Function on for a Fixed marker, in the case where a function was already on when the marker became Fixed. In this case the function value will be retained in the marker. It is also possible to have a Marker Function on for a Fixed marker in the case when the marker was off and was turned on as

Fixed

because

Delta

was pressed to create a reference marker - in which case the marker function, marker function width, Y Axis value and marker function result that the

Delta

marker had when

Delta

was pressed are copied into the Fixed marker. If

Delta

is pressed again, causing the fixed reference marker to move to the delta marker’s position, the marker function, marker function width, Y

Axis value and marker function result that the

Delta

marker had when

Delta

was pressed are again copied into the fixed reference marker.

If a Marker Function is on for a Fixed marker, the marker’s reported value is derived by the function.

Therefore you cannot directly set the X or Y value of a Fixed marker which has a marker function turned on. Indirect setting as detailed above or when a Peak Search is performed is allowed, as the Fixed marker is always placed on a trace and can derive its function value from the trace at the moment when it is placed.

Interval Markers

What is an interval marker? The band power marker computes the total power within a span in a nonzero span. The results computation must include the RBW. The interval power marker measures the average power across some time interval in zero span.

133

Common Measurement Functions 1

Interval Density is defined to be Interval Power divided by Bn. Bn is the noise bandwidth of the RBW filter, as noted and used within the Band Power computation.

Select Marker

See “Select Marker ” on page 1574 .

Marker Noise

Turns on the Marker Noise function for the selected marker, making it a noise marker. If the selected marker is off, it is turned on in

Normal

mode and located at the center of the screen.

When

Marker Noise

is selected while in the

Marker Function Off

state, the

Band Span

or

Interval

Span

is initialized to 5% of the screen width.

When

Marker Noise

is on, the marker’s Y Axis Result is the average noise level, normalized to a 1 Hz noise power bandwidth, in the band specified under the

Band Adjust

key.

See

“More Information” on page 134 .

See

“Off-trace Markers” on page 135

.

Key Path

Example

Notes

Dependencies

Couplings

Initial S/W Revision

Marker Fctn

CALC:MARK:FUNC NOIS turns on marker 1 as a noise marker.

CALC:MARK:FUNC? returns the current marker function for the marker specified. In this case it returns the string: NOIS.

CALC:MARK:Y? returns the y-axis value of the Marker Noise function for marker 1 (if Marker Noise is ON for marker 1). Note that the delta value when the Y axis unit is Watt is the square of the delta value when the Y axis unit is

Volt. For example, when the percent ratio with Y axis unit in Volt is 0.2, the percent ratio with Y axis unit in Watt will be 0.22 = 0.04. When you read the value out remotely you have to know whether your Y Axis Unit is log (dB), linear (V or A), or power (W).

See the description under the “ “Marker Function” on page 132

key, above.

Fixed markers: It is not possible to change the Band Function for a Fixed marker; so all of the Band Function keys are grayed out for a Fixed marker.

Average detector and Power Averaging auto selected when Marker Noise on

If the selected (specified) marker is off, selecting Marker Noise via front panel or SCPI will turn the marker on.

Prior to A.02.00

More Information

To guarantee accurate data for noise-like signals, a correction for equivalent noise bandwidth is made by the analyzer. The

Marker Noise

function accuracy is best when the detector is set to Average or Sample, because neither of these detectors will peak-bias the noise. The trade off between sweep time and variance of the result is best when Average Type is set to Power Averaging. Therefore, Auto coupling chooses the Average detector and Power Averaging when Marker Noise is on. Though the Marker Noise function works with all settings of detector and Average Type, using the positive or negative peak

134

Common Measurement Functions 1

detector gives less accurate measurement results.

Off-trace Markers

If a

Normal

or

Delta

noise marker is so near to the left or right edge of the trace that some of the band is off the trace, then it uses only that subset of the Band Width that is on-trace. If the marker itself is off-trace, its value becomes undefined.

Neither band/interval power nor band/interval density markers are defined if any part of the band is off-trace (unless they are Fixed with a stored function value in them), except that when the edges of the bandwidth are trivially off-screen, due to mathematical limitations in the analyzer or in the controlling computer, the result will still be considered valid.

Band/Interval Power

Turns on the Band/Interval Power function for the selected marker. If the selected marker is off it is turned on in

Normal

marker and located at the center of the screen.

When

Band/Interval Power

is selected while in the

Marker Function Off

state, the

Band Span

or

Interval Span

is initialized to 5% of the screen width.

If the detector mode for the detector on the marker’s trace is set to Auto, the average detector is selected.

If the Average type is set to Auto, Power Averaging is selected. Other choices for the detector or Average type will usually cause measurement inaccuracy.

Key Path

Example

Notes

Dependencies

Couplings

Marker Fctn

CALC:MARK:FUNC BPOW turns on marker 1 as a band power marker.

CALC:MARK2:FUNC? returns the current setting of marker function for marker 2. In this case it returns the string: BPOW.

CALC:MARK:Y? returns the y-axis value of the Band Power function for marker 1. Note that the delta value when the Y axis unit is Watt is the square of the delta value when the Y axis unit is Volt. For example, when the percent ratio with Y axis unit in Volt is 0.2, the percent ratio with Y axis unit in Watt will be 0.22 = 0.04. When you read the value out remotely you have to know whether your Y Axis Unit is log (dB), linear (V or A), or power (W).

See the description under the “

“Marker Function” on page 132

key, above.

Fixed markers: It is not possible to change the Band Function for a Fixed marker; so all of the Band Function keys are grayed out for a Fixed marker.

If the detector mode for the detector on the marker’s trace is set to Auto, the average detector is selected. If the Average type is set to Auto, Power

Averaging is selected.

If the selected (specified) marker is off, selecting Band Power via front panel or SCPI will turn the marker on.

Prior to A.02.00

Initial S/W Revision

Band/Interval Density

Turns on the Band/Interval Density function for the selected marker. If the selected marker is off it is

135

Common Measurement Functions 1

turned on in

Normal

marker mode and located at the center of the screen.

When

Band/Interval Density

is selected while in the

Marker Function Off

state, the

Band Span

or

Interval Span

is initialized to 5% of the screen width.

See

“More Information” on page 136 .

See

“What is band/interval density? ” on page 137

Key Path

Example

Notes

Notes

Dependencies

Couplings

State Saved

Initial S/W Revision

Marker Fctn

CALC:MARK:FUNC BDEN turns on marker 1 as a band density marker.

CALC:MARK:FUNC? returns the current setting of band function for the marker specified. In this case it returns the string: BDEN.

CALC:MARK:Y? returns the y-axis value of the Band Density function for marker 1. Note that the delta value when the Y axis unit is Watt is the square of the delta value when the Y axis unit is Volt. For example, when the percent ratio with Y axis unit in Volt is 0.2, the percent ratio with Y axis unit in Watt will be 0.22 = 0.04. When you read the value out remotely you have to know whether your Y Axis Unit is log (dB), linear (V or A), or power (W).

The zero-width case is treated as one bucket wide although it shows a width of

0.

When the trace the marker is on crosses domains, the width crosses domains as well, to remain the same percentage of the trace

See the description under the “ “Marker Function” on page 132

key, above.

Fixed markers: It is not possible to change the Band Function for a Fixed marker; so all of the Band Function keys are grayed out for a Fixed marker.

If the detector mode for the detector on the marker’s trace is set to Auto, the average detector is selected. If the Average type is set to Auto, Power

Averaging is selected.

If the selected (specified) marker is off, selecting Band Density via front panel or SCPI will turn the marker on.

n/a.

Prior to A.02.00

More Information

It may seem like the band density marker function is exactly like a function of a noise marker with variable width. But they are somewhat different. The Noise markers assume that the signal to be measured is noise-like. Based on this assumption, we can actually make reasonable measurements under very nonideal conditions: any detector may be used, any averaging type, any VBW. In contrast, the Band

Power and Band Density markers make no assumption about the statistics of the signal.

If the detector mode for the detector on the marker’s trace is set to Auto, the average detector is selected.

If the Average type is set to Auto, Power Averaging is selected. Other choices for the detector or Average type will usually cause measurement inaccuracy.

136

Common Measurement Functions 1

What is band/interval density?

On frequency domain traces, the average density across a band is the total band power divided by the bandwidth over which it is measured.

On time domain traces, interval density is the average power in the interval divided by the noise bandwidth of the RBW of the trace.

Off

Turns off band functions for the selected marker.

Key Path

Example

Notes

Dependencies

Couplings

Marker, Marker Fctn

:CALC:MARK:FUNC OFF turns off marker functions for marker 1

See the description under the “Marker” on page 1571 key.

Fixed markers: It is not possible to change the Band Function for a Fixed marker; so all of the Band Function keys are grayed out for a Fixed marker, including Off

Turning off the marker function has no effect on the band span nor does it turn the marker off.

Prior to A.02.00

Initial S/W Revision

Band Adjust

Opens a menu that lets you set the width or left or right edges of the band.

It is legal to change the width of the band even if there is no marker function on. Generally this can only happen by sending the SCPI command since access to the menu is restricted if no marker function is on.

Key Path

Dependencies

Couplings

Marker Function

If the marker is Fixed, Band Adjust is grayed out.

If the marker function is Off, Band Adjust is grayed out.

If any of the Band Adjust functions are the active function, the wings and arms of the selected marker display in green; otherwise they display in white.

Prior to A.02.00

Initial S/W Revision

Band/Interval Span

Sets the width of the span for the selected marker.

It is legal to change the width of the band even if there is no marker function on. Generally this can only happen by sending the SCPI command since access to the menu is restricted if no marker function is on.

In the table below, sweep_width = max(1,sweep_points–1) and sweep_points is the number of sweep points, set in the

Sweep

menu.

Key Path

Marker Fctn, Band Adjust

137

Common Measurement Functions 1

Remote Command

Example

Notes

Notes

Couplings

Preset

State Saved

Min

Max

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:

BAND:SPAN <freq>

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:

BAND:SPAN?

:CALC:MARK12:FUNC:BAND:SPAN 20 MHz sets the band span of marker

12 to 20 MHz

:CALC:MARK:FUNC:BAND:SPAN? queries the band span of Marker 1

Units are those of the trace’s domain, Hz for frequency domain, s for time domain.

Sending this command selects the subopcoded marker

The unit of the parameter must match the current domain of the trace the selected marker is on, or an invalid suffix error will be generated. If no unit is sent the fundamental unit for the trace domain will be used (Hz for freq domain traces, s for time domain traces).

Note that all the values provided in this table are only valid for frequency domain traces. If the current domain of the trace is time domain, values and unit will be different. In frequency domain, the Preset value is dependant on the frequency range of the instrument. The default value 1.3245 GHz is appropriate only if the instrument is a 26.5 GHz instrument (Option 526). In a

26.5 GHz Instrument, the default span is 26.49 GHz, so 5% of the span corresponds to 1.3245 GHz.

Changing the Band/Interval Span necessarily changes the Band/Interval Left and Band/Interval Right values

Band/Interval Span is set to 0 when the marker is turned off

Band/Interval Span is set to 5% of span when any marker function is turned on if and only if it is zero at that time

If 0, set to 5% of span, when a marker function is turned on

Saved in Instrument State

0 Hz

Infinity. Unlike legacy analyzers, where the markers were forced to be on screen, X-Series marker values are not limited and do not clip

:CALCulate:MARKer[1]|2|3|4:X:SPAN

Prior to A.02.00

Backwards Compatibility SCPI

Initial S/W Revision

Band/Interval Left

Sets the left edge frequency or time for the band of the selected marker. The right edge is unaffected.

It is legal to change the width of the band even if there is no marker function on. Generally this can only happen by sending the SCPI command since access to the menu is restricted if no marker function is on.

In the table below, sweep_width = max(1,sweep_points–1) and sweep_points is the number of sweep

138

Common Measurement Functions 1

points, set in the

Sweep

menu.

Key Path

Remote Command

Marker Fctn, Band Adjust

Example

Notes

Notes

Couplings

Preset

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:

BAND:LEFT <freq>

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:

BAND:LEFT?

:CALC:MARK12:FUNC:BAND:LEFT 20 GHz sets the left edge of the band span of marker 12 to 20 GHz

:CALC:MARK:FUNC:BAND:LEFT? queries the band span of Marker 1

Units are those of the trace’s domain, Hz for frequency domain, s for time domain. When the left edge is moved, the right edge stays anchored; thus, the marker’s frequency will change.

Sending this command selects the subopcoded marker

The unit of the parameter must match the current domain of the trace the selected marker is on, or an invalid suffix error will be generated.If no unit is sent the fundamental unit for the trace domain will be used (Hz for freq domain traces, s for time domain traces).

Note that all the values provided in this table are only valid for frequency domain traces. If the current domain of the trace is time domain, values and unit will be different. In frequency domain, the Preset value is dependant on the frequency range of the instrument. The default value 1.3245 GHz is appropriate only if the instrument is a 26.5 GHz instrument (Option 526). In a

26.5 GHz Instrument, the default span is 26.49 GHz, so 5% of the span corresponds to 1.3245 GHz.

Changing the Band/Interval Left necessarily changes the Band/Interval Span and Band/Interval Center values

Band/Interval Span is set to 0 when the marker is turned off so that means

Band/Interval Left is set to the center value at this time

Band/Interval Span is set to 5% of span when any marker function is turned on if and only if it is zero at that time

If 0, Band/Interval Span is set to 5% of span, when a marker function is turned on, which affects Band/Interval Left

State Saved

Min

Saved in Instrument State

0 Hz

Max Infinity. Unlike legacy analyzers, where the markers were forced to be on screen, X-Series marker values are not limited and do not clip

Backwards Compatibility SCPI :CALCulate:MARKer[1]|2|3|4:X:STARt

(This legacy command was used to control the Reference marker in Delta

Pair/Band Pair mode, and is aliased to the new command.)

Initial S/W Revision Prior to A.02.00

139

Common Measurement Functions 1

Band/Interval Right

Sets the right edge frequency or time for the band of the selected marker. The left edge is unaffected

In the table below, sweep_width = max(1,sweep_points–1) and sweep_points is the number of sweep points, set in the

Sweep

menu.

It is legal to change the width of the band even if there is no marker function on. Generally this can only happen by sending the SCPI command since access to the menu is restricted if no marker function is on.

Key Path

Remote Command

Example

Notes

Notes

Couplings

Preset

State Saved

Min

Max

Marker Fctn, Band Adjust

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:

BAND:RIGHt <freq>

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:

BAND:RIGHt?

:CALC:MARK12:FUNC:BAND:RIGHt 20 GHz sets the right edge of the band span of marker 12 to 20 GHz

:CALC:MARK:FUNC:BAND:RIGHt? queries the band span of Marker 1

Units are those of the trace’s domain, Hz for frequency domain, s for time domain. When the right edge is moved, the left edge stays anchored; thus, the marker’s frequency will change.

Sending this command selects the subopcoded marker

The unit of the parameter must match the current domain of the trace the selected marker is on, or an invalid suffix error will be generated. If no unit is sent the fundamental unit for the trace domain will be used (Hz for freq domain traces, s for time domain traces).

Note that all the values provided in this table are only valid for frequency domain traces. If the current domain of the trace is time domain, values and unit will be different. In frequency domain, the Preset value is dependant on the frequency range of the instrument. The default value 1.3245 GHz is appropriate only if the instrument is a 26.5 GHz instrument (Option 526). In a

26.5 GHz Instrument, the default span is 26.49 GHz, so 5% of the span corresponds to 1.3245 GHz.

Changing the Band/Interval Right necessarily changes the Band/Interval Span and Band/Interval Center values

Band/Interval Span is set to 5% of span when any marker function is turned on if and only if it is zero at that time

If 0, Band/Interval Span is set to 5% of span, when a marker function is turned on, which affects Band/Interval Right

Saved in Instrument State

0 Hz

Infinity. Unlike legacy analyzers, where the markers were forced to be on screen, X-Series marker values are not limited and do not clip

140

Common Measurement Functions 1

Backwards Compatibility SCPI :CALCulate:MARKer[1]|2|3|4:X:STOP

(This legacy command was used to control the Delta marker in Delta

Pair/Band Pair mode, and is aliased to the new command. For compatibility.

Note that if you were using the old command for Band Power measurements it will work just fine.)

Initial S/W Revision Prior to A.02.00

Measure at Marker

This key and all the keys in this menu only appear with Option EMC installed and licensed.

This key opens up a menu which contains the Measure at Marker functions. This key only appears with option EMC installed.

Key Path

Initial S/W Revision

Marker Function

A.02.00

Measure at Marker

When this key is pressed, the analyzer executes one Measure at Marker function and then returns.

Measure at Marker goes to the frequency of the selected marker and takes a reading with each of the three detectors selected in the Detectors menu, using the dwell times specified there, then displays the readings in a window on the display, using the current Y-Axis Unit.

When the Measure at Marker is complete, the analyzer restores all settings to their pre-Measure-at-Marker values and normal sweeps resume.

Key Path

Remote Command

Example

Marker Function, Measure at Marker

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:

MAMarker?

:CALC:MARK2:FUNC:MAM?

Performs a Measure at Marker function at Marker 2’s current frequency and, when completed, returns the results of the measure at marker window in a query

141

Common Measurement Functions 1

Notes

Dependencies

Couplings

Status Bits/OPC dependencies

Backwards Compatibility SCPI

This query command returns comma separated values for the 3 specified detectors and the frequency value of the marker. If a Detector is off or if no measurement has yet completed, –999.0 will be returned. This can happen, for example, if you are operating with too large a value of (span/sweep points) and the Measure at Marker function does not execute but instead puts up the advisory message, “Span per point too large, narrow span or increase RBW or number of points” (see below).

The size of the return data array is fixed at 4. The elements are:

1. Detector 1 value ( if off, –999.0 for backwards compatibility)

2. Detector 2 value ( if off, –999.0 for backwards compatibility)

3. Detector 3 value ( if off, –999.0 for backwards compatibility)

4. Frequency of Marker

If a sweep is in process when this function executes it aborts, and restarts after the function is complete.

This command is not backwards compatible with the E7400 and PSA option

239 so the Backwards Compatibility command is included.

If

BW & Avg Type

is in Autocoupled state, the (up to three) measurements taken by Measure at Marker are taken with Auto Coupled settings for the functions in the BW menu, even if those functions are in manual.

If the specified Marker is not on, the analyzer turns it on at center of screen and does a peak search before performing the function.

OPC goes true when the measurement is complete

:MEASure:EMI:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12?

(Performs a Measure at Marker function at specified marker’s current frequency and returns the results)

A.02.00

Initial S/W Revision

Measure at Marker presents its information in a separate window which normally appears in the upper right of the display but can be repositioned to the upper left.

142

Common Measurement Functions 1

The Measure at Marker box shows the detector name for the selected detectors and “Off” for those not selected. The names used are:

Name

Normal

Detector

Normal

Peak

Sample

Neg Peak

RMS

Peak

Sample

Negative Peak

Average detector with Power Average (RMS)

Log Avg

VoltageAvg

Quasi Peak

EMI Avg

Average detector with Log-Pwr Average

Average detector with Voltage Average

Quasi Peak

EMI Average

RMS Avg RMS Average

The marker frequency is shown in the “Freq” field. The measured value is shown for all detectors except those that are “Off.” For these, --- is displayed. The current Y-Axis unit is used, and the precision that is used for the detector value displays is exactly the same as for the Marker The precision used for the

Frequency display is six significant digits.

The sequence of steps in the measurement is as follows:

143

Common Measurement Functions 1

• Any sweep in progress is aborted.

• If in Zero Span, the Center Frequency is used as the frequency at which to take the reading, since in

Zero Span, all markers are by definition at the Center Frequency

• If not in Zero Span:

— If the selected marker is Off, it is first turned on in the center of the screen and a peak search performed.

— If the selected marker is on, but offscreen, it is first moved to the center of the screen and a peak search performed. .

— A frequency “zoom” function is performed to determine the frequency of the selected marker to the required precision. If you are operating with too large a value of (span/sweep points) then the

Measure at Marker window will not display, but instead an advisory message, “Span per point too large, narrow span or increase RBW or number of points”.This means you have chosen a combination of RBW, span and sweep points that makes each trace point much wider than the

RBW, so that the trace point in which the signal appears is an inadequately precise measure of its frequency—for example, with a 30 MHz to 1000 MHz span, 601 trace points and 120 kHz RBW, each trace point is 13 times as wide as the RBW. In this case, a SCPI query of the results will yield –999 dBm for each detector.

— If the zoom is successful, the analyzer goes to zero span at this frequency

• Each detector is then read in successive single-point zero span sweeps, using a sweep time equal to the specified dwell time. The value displayed by Measure at Marker represents the maximum value output by the detector during the dwell timeAutocoupled bandwidth and average type settings are used for each detector unless the

BW & Avg Type

key is set to

As Set

, in which case the current bandwidth and average type settings are used.

• Each result is then displayed in the measure at marker window as it becomes available.

• The analyzer returns to its pre-Measure at Marker span and settings after executing a Measure at

Marker function, including Bandwidth, Avg Type, and EMC Std - regardless of the setting of

BW &

Avg Type

• Finally, if the sweep had to be aborted, the aborted sweep is restarted.

While the function is executing, all the fields except Freq show --- for their values until the measurement is complete for that detector. As each detector is read, an informational message is displayed in the status line, for example,

Measuring with detector 1 (Peak) with RBW=120 kHz

After the last detector, the status line is cleared.

Meas at Marker Window

This key opens a menu which controls the Measure at Marker window.

Key Path

Readback

Marker Function, Measure at Marker

In square brackets, the state of the window then the window position, separated by commas, as [On, Left]

144

Common Measurement Functions 1

Initial S/W Revision A.02.00

Window On/Off

This key turns the Measure at Marker window on and off. It turns on automatically when Measure at Marker is initiated and turns off on a Preset. If the Window is turned on without a Measure at Marker result, --- is displayed for each result for which the detector is not “Off”.

Key Path

Remote Command

Example

Couplings

Preset

State Saved

Readback Text

Initial S/W Revision

Marker Function, Measure at Marker, Meas at Marker Window

:DISPlay:WINDow:MAMarker[:STATe] ON|OFF|1|0

:DISPlay:WINDow:MAMarker[:STATe]?

:DISP:WIND:MAM ON

The window turns on automatically when Measure at Marker is initiated and turns off on a Preset.

Off

Saved in state

On|Off

A.02.00

Position Left/Right

This key controls the placement of the Measure at Marker window on the display.

Key Path

Remote Command

Example

Preset

State Saved

Readback Text

Initial S/W Revision

Marker Function, Measure at Marker, Meas at Marker Window

:DISPlay:WINDow:MAMarker:POSition LEFT|RIGHt

:DISPlay:WINDow:MAMarker:POSition?

:DISP:WIND:MAM:POS RIGH

Right

Saved in state

Left|Right

A.02.00

Detectors

This key opens up a menu that allows you to configure the detectors to be used for the Measure at

Marker reading. Any of the analyzer’s detectors can be used for each of the three detectors, or any of the three can be turned off. The dwell time for each detector is also settable.

When performing a Meas at Marker, the dwell time settings that you select will depend on the characteristics of the emission you are measuring. The default dwell time (200 ms) should work well for typical EUT emissions, but sometimes you will encounter emissions for which the defaults are not optimal. This is especially the case for emissions that vary slowly over time or have a slow repetition rate. By lengthening the dwell times you can increase the likelihood of accurately measuring these low

145

Common Measurement Functions 1

repetition rate signals.

When Measure at marker is activated, the receiver makes a zero span measurement for each of the (up to) three detectors selected, using the Dwell Time set for each detector. If the signal's repetition period is greater than 200 ms (the default setting), the dwell time should be increased to capture at least two and preferably more repetitions of the signal. Additionally, if you do not need or do not wish to use a detector to make a measurement, that specific detector may be turned off.

If the Measure at Marker window is being displayed, and one of the detectors is changed, any value being displayed for that detector changes to “---“ until the next successful reading from that detector.

Key Path

Remote Command

Example

State Saved

Initial S/W Revision

Marker Function, Measure at Marker,

:CALCulate:MAMarker:DETector[1]|2|3

OFF|NORMal|AVERage|POSitive|SAMPle|NEGative|QPEak|EAVer age|RAVerage

:CALCulate:MAMarker:DETector[1]|2|3?

:CALC:MAM:DET2 QPE

Sets the detector for measure at marker detector 2 to Quasi peak

:CALC:MAM:DET OFF

Sets the detector for measure at marker detector 1 to Off

Saved in state

A.02.00

Key Path

Remote Command

Example

State Saved

Backwards Compatibility SCPI

Initial S/W Revision

Marker Function, Measure at Marker,

:CALCulate:MAMarker:DETector[1]|2|3:DWELl <dwell time>

:CALCulate:MAMarker:DETector[1]|2|3:DWELl?

:CALC:MAM:DET2:DWEL 500 ms

Sets the detector for measure at marker detector 2 to dwell for 500 ms

Saved in state

[:SENSe]:EMI:MEASure:DETector:DWELl <dwell time>

! Sets all of the detectors dwell time to the specified amount

A.02.00

Detector 1

This menu lets you select the detector to be used for Detector 1, or turn Detector 1 off. This is a 1-of-N menu that shows the normal list of detectors, but with the “Auto” key replaced by “Off”.

Key Path

Remote Command

Marker Function, Measure at Marker, Detectors

See “Detectors” on page 145

.

146

Common Measurement Functions 1

Example

Preset

State Saved

Readback Text

Initial S/W Revision

:CALC:MAM:DET QPE

Sets the detector for measure at marker detector 1 to Quasi peak

:CALC:MAM:DET OFF

Sets the detector for measure at marker detector 1 to Off

Peak

Saved in state

Detector name

A.02.00

Detector 2

This menu lets you select the detector to be used for Detector 2, or turn Detector 2 off. This is a 1-of-N menu that shows the normal list of detectors, but with the “Auto” key replaced by “Off”.

Key Path

Remote Command

Example

Marker Function, Measure at Marker, Detectors

See

“Detectors” on page 145 .

:CALC:MAM:DET2 QPE

Sets the detector for measure at marker detector 2 to Quasi peak

:CALC:MAM:DET2 OFF

Sets the detector for measure at marker detector 2 to Off

Preset

State Saved

Quasi Peak

Saved in state

Readback Text Detector name

Backwards Compatibility SCPI [:SENSe]:EMI:MEASure:DETector:QPEak[:STATe] OFF|ON|0|1|

! If sent with On as a parameter, sets detector 2 to Quasi Peak

! If sent with Off as a parameter, sets detector 2 to Off

Initial S/W Revision A.02.00

Detector 3

This menu lets you select the detector to be used for Detector 3, or turn Detector 3 off. This is a 1-of-N menu that shows the normal list of detectors, but with the “Auto” key replaced by “Off”.

Key Path

Remote Command

Marker Function, Measure at Marker, Detectors

See

“Detectors” on page 145 .

147

Common Measurement Functions 1

Example

Preset

State Saved

Readback Text

Backwards Compatibility SCPI

:CALC:MAM:DET3 QPE

Sets the detector for measure at marker detector 1 to Quasi peak

:CALC:MAM:DET3 OFF

Sets the detector for measure at marker detector 1 to Off

EMI Average

Saved in state

Detector name

[:SENSe]:EMI:MEASure:DETector:AVERage[:STATe] OFF|ON|0|1|

! If sent with On as a parameter, sets detector 3 to EMI Average

! If sent with Off as a parameter, sets detector 3 to Off

A.02.00

Initial S/W Revision

Detector 1 Dwell Time

This is the time specified by the user to dwell while taking the measurement for detector 1. The minimum allowed dwell time is based on the current detector If “Off” is selected for detector 1, this key is grayed out and shows 200 ms.

Key Path

Remote Command

Example

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

Marker Function, Measure at Marker, Detectors

See “Detectors” on page 145

.

:CALC:MAM:DET:DWEL 400 ms

Sets the dwell time for detector 1 to 400 ms

200 ms

Saved in state

1 ms

60 s s

A.02.00

Detector 2 Dwell Time

This is the time specified by the user to dwell while taking the measurement for detector 2. The minimum allowed dwell time is based on the current detector. If “Off” is selected for detector 2, this key is grayed out and shows

200 ms.

Key Path

Remote Command

Example

Marker Function, Measure at Marker, Detectors

See “Detectors” on page 145

.

:CALC:MAM:DET2:DWEL 400 ms

Sets the dwell time for detector 2 to 400 ms

148

Common Measurement Functions 1

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

200 ms

Saved in state

1 ms

60 s s

A.02.00

Detector 3 Dwell Time

This is the time specified by the user to dwell while taking the measurement for detector 3. The minimum allowed dwell time is based on the current detector. If “Off” is selected for detector 3, this key is grayed out and shows 200 ms.

Key Path

Remote Command

Example

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

Marker Function, Measure at Marker, Detectors

See

“Detectors” on page 145 .

:CALC:MAM:DET3:DWEL 400 ms

Sets the dwell time for detector 1 to 400 ms

200 ms

Saved in state

1 ms

60 s s

A.02.00

BW & Avg Type

This key controls the type of bandwidth and average type coupling used in Measure at Marker.

If set to “Autocoupled”, then the RBW and Average Type are selected by the instrument during the

Measure at Marker function, according to the normal Autocouple rules, regardless of whether RBW and

Average Type are currently in Auto. If set to “As Set”, then the current value for RBW and Average

Type are used (which of course, could also be “Auto”).

Here are the details of the two modes:

If

BW & Avg Type

is set to

Autocoupled

,

Measure at Marker

behaves as follows:

1) The

EMC Std

changes to CISPR if any of the CISPR detectors (EMI Avg, RMS Avg, QPD) becomes selected; for all other detectors, the value of

EMC Std

that existed before Measure at Marker is used.

2)

RBW

autocouples throughout Measure at Marker, even if

RBW

is set to

Manual.

The autocouple rules are based on whatever the instantaneous setting of EMC Std, Span, and Center Freq are.

If

BW & Avg Type

is set to

As Set

,

Measure at Marker

behaves as follows:

149

Common Measurement Functions 1

1) The

EMC Std

never changes; so if it is set to

None

it stays at

None

throughout, even if one of the

CISPR detectors is selected.

2) If

RBW

is set to

Auto

, then

RBW

autocouples throughout Measure at Marker. The autocouple rules are based on whatever the setting of EMC Std, Span, and Center Freq are.

3) If

RBW

is set to

Manual

, the RBW never changes at all throughout Meaure at Marker, it stays at the value to which it was set before Measure at Marker began.

The analyzer returns to its pre-Measure at Marker span and settings after executing a Measure at Marker function, including Bandwidth, Avg Type, and EMC Std.

It is important to note that, when RBW is coupled to Frequency, as it is when

EMC Std

is anything but

“None”, for all EMI measurements, the frequency it is coupled to for Measure at Marker is the

MARKER frequency, not the Center Frequency.

Key Path

Remote Command

Example

Preset

State Saved

Readback Text

Initial S/W Revision

Marker Function, Measure at Marker

:CALCulate:MAMarker:COUPling ON|OFF|1|0

:CALCulate:MAMarker:COUPling?

:CALC:MAM:COUP ON

Autocoupled

Saved in state

Autocoupled|As Set

A.02.00

Center Presel On/Off

This key controls the automatic centering of the preselector for the Measure at Marker function.

When Center Presel is On, the first step in performing the Measure at Marker function is to perform a

Presel Center. This is not performed if the microwave preselector is off, or the selected marker’s frequency is below Band 1. If the function is not performed, no message is generated.

Remote Command

Example

Initial S/W Revision

Key Path

Dependencies

Preset

:CALCulate:MAMarker:PCENter ON|OFF|1|0

:CALCulate:MAMarker:PCENter?

:CALC:MAM:PCEN ON

A.02.00

Marker Function, Measure at Marker

Blank in models that do not include a preselector, such as option 503. If the

SCPI is sent in these instruments, it is accepted without error, and the query always returns 0.

On

150

Common Measurement Functions 1

Backwards Compatibility SCPI [:SENSe]:EMI:MEASure:PCENter[:STATe] OFF|ON|0|1

[:SENSe]:EMI:MEASure:PCENter[:STATe]?

Marker To

The Marker -> key accesses menu keys that can copy the current marker value into other instrument parameters (for example, Center Freq). The currently selected marker is made the active function on entry to this menu (if the currently selected marker is not on when you press this front panel key, it will be turned on at the center of the screen as a normal type marker and then made the active function).

The

Marker ->

(or Marker To) feature is used to quickly assign a marker’s x- or y-axis value to another parameter. For example, if a marker’s x-axis value is 500 MHz and y-axis value is –20 dBm, pressing

Mkr -> CF

would assign 500 MHz to

Center Freq

and pressing

Mkr - >Ref Lvl

would assign –20 dBm to

Ref Level

.

Notes

Initial S/W Revision

All Marker To functions executed from the front panel use the selected marker’s values, while all Marker To remote commands specify in the command which marker’s value to use.

Consistent with other remote marker commands, sending a Marker To remote command will never change which marker is selected.

Prior to A.02.00

Mkr->CF

Sets the center frequency of the analyzer to the frequency of the selected marker. The marker stays at this frequency, so it moves to the center of the display. In delta marker mode, this function sets the center frequency to the x-axis value of the delta marker. When the frequency scale is in log mode, the center frequency is not at the center of the display.

If the currently selected marker is not on when this key is pressed, it will be turned on at the center of the screen as a normal type marker.

Key Path

Remote Command

Example

Notes

Dependencies

Couplings

Initial S/W Revision

Marker ->

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:CEN

Ter

CALC:MARK2:CENT sets the CF of the analyzer to the value of marker 2.

Sending this command selects the subopcoded marker

If specified marker is off, this command will turn it on at the center of the screen as a normal type marker.

This function is not available (key is grayed out) when x-axis is the time domain

All the usual couplings associated with setting Center Frequency apply (see the Frequency Section).

Prior to A.02.00

151

Common Measurement Functions 1

Mkr->CF Step

Sets the center frequency (CF) step size of the analyzer to the marker frequency, or in a delta-marker mode, to the frequency difference between the delta and reference markers.

If the currently selected marker is not on when this key is pressed, it will be turned on at the center of the screen as a normal type marker.

Key Path

Remote Command

Example

Notes

Dependencies

Couplings

Marker ->

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:STE

P

CALC:MARK1:STEP sets the CF step to the value (or delta value) of marker

1.

Sending this command selects the subopcoded marker

If specified marker is off, this command will turn it on at the center of the screen as a normal type marker.

This function is not available (key is grayed out) when x-axis is the time domain

All the usual couplings associated with setting CF Step apply (see the

Frequency Section).

Prior to A.02.00

Initial S/W Revision

Mkr->Start

Changes the start frequency to the frequency of the selected marker. The marker stays at this frequency, so it moves to the left edge of the display. In delta marker mode, this function sets the start frequency to the x-axis value of the delta marker.

If the currently selected marker is not on when this key is pressed, it will be turned on at the center of the screen as a normal type marker.

Key Path

Remote Command

Example

Notes

Dependencies

Couplings

Marker ->

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:STA

Rt

CALC:MARK1:STAR sets the start frequency to the value (or delta value) of marker 1.

Sending this command selects the subopcoded marker

If specified marker is off, this command will turn it on at the center of the screen as a normal type marker.

This function is not available (key is grayed out) when x-axis is the time domain

All the usual couplings associated with setting Start Frequency apply (see the

Frequency Section).

152

Common Measurement Functions 1

Initial S/W Revision Prior to A.02.00

Mkr->Stop

Changes the stop frequency to the frequency of the selected marker. The marker stays at this frequency, so it moves to the right edge of the display. In delta marker mode, this function sets the stop frequency to the x-axis value of the delta marker.

If the currently selected marker is not on when this key is pressed, it will be turned on at the center of the screen as a normal type marker.

Key Path

Remote Command

Example

Notes

Dependencies

Couplings

Marker ->

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:STO

P

CALC:MARK3:STOP sets the stop frequency to the value (or delta value) of marker 3.

Sending this command selects the subopcoded marker

If specified marker is off, this command will turn it on at the center of the screen as a normal type marker.

This function is not available (key is grayed out) when x-axis is the time domain

All the usual couplings associated with setting Stop Frequency apply (see the

Frequency Section).

Prior to A.02.00

Initial S/W Revision

Mkr

Δ->Span

Sets the start and stop frequencies to the values of the delta markers. That is, it moves the lower of the two marker frequencies to the start frequency and the higher of the two marker frequencies to the stop frequency. The marker mode is unchanged and the two markers (delta and reference) end up on opposite edges of the display.

Key Path

Remote Command

Example

Notes

Dependencies

Couplings

Marker ->

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:DEL

Ta:SPAN

CALC:MARK2:DELT:SPAN sets the start and stop frequencies to the values of marker 2 and its reference marker.

Sending this command selects the subopcoded marker

This function is only available when the selected marker is a delta marker.

Otherwise the key is grayed out.

In addition, this function is not available when x-axis is the time domain

All the usual couplings associated with setting Span apply (see the Section

“ “SPAN X Scale” on page 1473 ”).

153

Common Measurement Functions 1

Backwards Compatibility SCPI :CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:SPAN

Initial S/W Revision Prior to A.02.00

Mkr

Δ->CF

Sets the center frequency to the frequency difference between the selected marker and its reference marker. The marker is then changed to a Normal marker and placed at the center of span.

Key Path

Remote Command

Example

Notes

Dependencies

Marker ->

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:DEL

Ta:CENTer

CALC:MARK2:CENT sets the CF of the analyzer to the value of marker 2.

Sending this command selects the subopcoded marker

This function is only available when the selected marker is a delta marker.

Otherwise the key is grayed out.

In addition, this function is not available when x-axis is the time domain

Prior to A.02.00

Initial S/W Revision

Mkr->Ref Lvl

Sets the reference level to the amplitude value of the selected marker, moving the marked point to the reference level (top line of the graticule). The marker’s mode (Normal, Delta, Fixed) doesn’t matter in this case. For example, given a delta marker, if the delta marker is the selected marker, its amplitude is applied to the reference level. If the reference marker is selected, its amplitude is applied to the reference level.

If the currently selected marker is not on when this key is pressed, it will be turned on at the center of the screen as a normal type marker, and its amplitude applied to the reference level.

Key Path

Remote Command

Example

Notes

Couplings

Initial S/W Revision

Marker ->

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:SET]:RLE

Vel

CALC:MARK2:RLEV sets the reference level of the analyzer to the amplitude of marker 2.

Sending this command selects the subopcoded marker

If specified marker is off, this command will turn it on at the center of the screen as a normal type marker.

All the usual couplings associated with setting Reference Level apply.

Prior to A.02.00

154

Common Measurement Functions 1

Meas

The information in this section is common to all measurements. For key and remote command information for a specific measurement, refer to the section that describes the measurement of interest.

Measurements available under the Meas key are specific to the current Mode.

When viewing Help for measurements, note the following:

NOTE

Operation for some keys differs between measurements. The information displayed in Help pertains to the current measurement. To see how a key operates in a different measurement, exit Help (press the Cancel Esc key), select the measurement, then reenter Help (press the Help key) and press that key.

Key Path

Initial S/W Revision

Front-panel key

Prior to A.02.00

Remote Measurement Functions

This section contains the following topics:

“Measurement Group of Commands” on page 156

“Current Measurement Query (Remote Command Only) ” on page 158

“Limit Test Current Results (Remote Command Only)” on page 159

“Data Query (Remote Command Only)” on page 159

“Calculate/Compress Trace Data Query (Remote Command Only)” on page 159

“Calculate Peaks of Trace Data (Remote Command Only)” on page 164

“Format Data: Numeric Data (Remote Command Only)” on page 166

“Format Data: Byte Order (Remote Command Only)” on page 168

Initial S/W Revision Prior to A.02.00

155

Common Measurement Functions 1

Measurement Group of Commands

Measure Commands:

:MEASure:<measurement>[n]?

This is a fast single-command way to make a measurement using the factory default instrument settings. These are the settings and units that conform to the Mode Setup settings (e.g. radio standard) that you have currently selected.

• Stops the current measurement (if any) and sets up the instrument for the specified measurement using the factory defaults

• Initiates the data acquisition for the measurement

• Blocks other SCPI communication, waiting until the measurement is complete before returning results.

• If the function does averaging, it is turned on and the number of averages is set to 10.

• After the data is valid it returns the scalar results, or the trace data, for the specified measurement. The type of data returned may be defined by an [n] value that is sent with the command.

The scalar measurement results will be returned if the optional [n] value is not included, or is set to 1. If the

[n] value is set to a value other than 1, the selected trace data results will be returned. See each command for details of what types of scalar results or trace data results are available.

ASCII is the default format for the data output. (Older versions of Spectrum Analysis and Phase Noise mode measurements only use ASCII.) The binary data formats should be used for handling large blocks of data since they are smaller and faster than the ASCII format. Refer to the FORMat:DATA command for more information.

If you need to change some of the measurement parameters from the factory default settings you can set up the measurement with the CONFigure command. Use the commands in the SENSe:<measurement> and

CALCulate:<measurement> subsystems to change the settings. Then you can use the READ? command to initiate the measurement and query the results.

If you need to repeatedly make a given measurement with settings other than the factory defaults, you can use the commands in the SENSe:<measurement> and CALCulate:<measurement> subsystems to set up the measurement. Then use the READ? command to initiate the measurement and query results.

Measurement settings persist if you initiate a different measurement and then return to a previous one. Use

READ:<measurement>? if you want to use those persistent settings. If you want to go back to the default settings, use MEASure:<measurement>?.

156

Common Measurement Functions 1

Configure Commands:

:CONFigure:<measurement>

This command stops the current measurement (if any) and sets up the instrument for the specified measurement using the factory default instrument settings. It does not initiate the taking of measurement data unless

INIT:CONTinuous is ON. If you change any measurement settings after using the CONFigure command, the

READ command can be used to initiate a measurement without changing the settings back to their defaults.

In the Swept SA measurement in Spectrum Analyzer mode the CONFigure command also turns the averaging function on and sets the number of averages to 10 for all measurements.

:CONFigure:NDEFault<measurement> stops the current measurement and changes to the specified measurement. It does not change the settings to the defaults. It does not initiate the taking of measurement data unless INIT:CONTinuous is ON.

The CONFigure? query returns the current measurement name.

The CONFigure:CATalog? query returns a quoted string of all measurement names in the current mode. For example, "SAN, CHP, OBW, ACP, PST, TXP, SPUR, SEM, LIST".

Fetch Commands:

:FETCh:<measurement>[n]?

This command puts selected data from the most recent measurement into the output buffer. Use FETCh if you have already made a good measurement and you want to return several types of data (different [n] values, for example, both scalars and trace data) from a single measurement. FETCh saves you the time of re-making the measurement. You can only FETCh results from the measurement that is currently active, it will not change to a different measurement. An error is reported if a measurement other than the current one, is specified.

If you need to get new measurement data, use the READ command, which is equivalent to an INITiate followed by a FETCh.

The scalar measurement results will be returned if the optional [n] value is not included, or is set to 1. If the [n] value is set to a value other than 1, the selected trace data results will be returned. See each command for details of what types of scalar results or trace data results are available. The binary data formats should be used for handling large blocks of data since they are smaller and transfer faster then the ASCII format. (FORMat:DATA)

FETCh may be used to return results other than those specified with the original READ or MEASure command that you sent.

INITiate Commands:

157

Common Measurement Functions 1

:INITiate:<measurement>

This command is not available for measurements in all the instrument modes:

• Initiates a trigger cycle for the specified measurement, but does not output any data. You must then use the

FETCh<meas> command to return data. If a measurement other than the current one is specified, the instrument will switch to that measurement and then initiate it.

• For example, suppose you have previously initiated the ACP measurement, but now you are running the channel power measurement. If you send INIT:ACP? it will change from channel power to ACP and will initiate an ACP measurement.

• Does not change any of the measurement settings. For example, if you have previously started the ACP measurement and you send INIT:ACP? it will initiate a new ACP measurement using the same instrument settings as the last time ACP was run.

• If your selected measurement is currently active (in the idle state) it triggers the measurement, assuming the trigger conditions are met. Then it completes one trigger cycle. Depending upon the measurement and the number of averages, there may be multiple data acquisitions, with multiple trigger events, for one full trigger cycle. It also holds off additional commands on GPIB until the acquisition is complete.

READ Commands:

:READ:<measurement>[n]?

• Does not preset the measurement to the factory default settings. For example, if you have previously initiated the ACP measurement and you send READ:ACP? it will initiate a new measurement using the same instrument settings.

• Initiates the measurement and puts valid data into the output buffer. If a measurement other than the current one is specified, the instrument will switch to that measurement before it initiates the measurement and returns results.

For example, suppose you have previously initiated the ACP measurement, but now you are running the channel power measurement. Then you send READ:ACP? It will change from channel power back to ACP and, using the previous ACP settings, will initiate the measurement and return results.

• Blocks other SCPI communication, waiting until the measurement is complete before returning the results

If the optional [n] value is not included, or is set to 1, the scalar measurement results will be returned. If the

[n] value is set to a value other than 1, the selected trace data results will be returned. See each command for details of what types of scalar results or trace data results are available. The binary data formats should be used when handling large blocks of data since they are smaller and faster then the ASCII format.

(FORMat:DATA)

Initial S/W Revision Prior to A.02.00

Current Measurement Query (Remote Command Only)

This command returns the name of the measurement that is currently running.

Remote Command:

Example:

:CONFigure?

CONF?

158

Common Measurement Functions 1

Initial S/W Revision: Prior to A.02.00

Limit Test Current Results (Remote Command Only)

Queries the status of the current measurement limit testing. It returns a 0 if the measured results pass when compared with the current limits. It returns a 1 if the measured results fail any limit tests.

Remote Command:

Example:

Initial S/W Revision:

:CALCulate:CLIMits:FAIL?

CALC:CLIM:FAIL? queries the current measurement to see if it fails the defined limits.

Returns a 0 or 1: 0 it passes, 1 it fails.

Prior to A.02.00

Data Query (Remote Command Only)

Returns the designated measurement data for the currently selected measurement and subopcode.

n = any valid subopcode for the current measurement. See the measurement command results table for your current measurement, for information about what data is returned for the subopcodes.

This command uses the data setting specified by the FORMat:BORDer and FORMat:DATA commands and can return real or ASCII data. (See the format command descriptions under Input/Output in the

Analyzer Setup section.)

Remote Command:

Notes:

:CALCulate:DATA[n]?

The return trace depends on the measurement.

In CALCulate:<meas>:DATA[n], n is any valid subopcode for the current measurement. It returns the same data as the FETCh:<measurement>? query where <measurement> is the current measurement.

Initial S/W Revision: Prior to A.02.00

Calculate/Compress Trace Data Query (Remote Command Only)

Returns compressed data for the currently selected measurement and subopcode [n].

n = any valid sub-opcode for that measurement. See the MEASure:<measurement>? command description of your specific measurement for information on the data that can be returned.

The data is returned in the current Y Axis Unit of the analyzer. The command is used with a sub-opcode

<n> (default=1) to specify the trace. With trace queries, it is best if the analyzer is not sweeping during the query. Therefore, it is generally advisable to be in Single Sweep, or Update=Off.

This command is used to compress or decimate a long trace to extract and return only the desired data. A typical example would be to acquire N frames of GSM data and return the mean power of the first burst

159

Common Measurement Functions 1

in each frame. The command can also be used to identify the best curve fit for the data.

Remote Command:

Notes:

:CALCulate:DATA<n>:COMPress?

BLOCk|CFIT|MAXimum|MINimum|MEAN|DMEan|RMS|RMSCubed|SAMP le|SDEViation|PPHase

[,<soffset>[,<length>[,<roffset>[,<rlimit>]]]]

The command supports 5 parameters. Note that the last 4

(<soffset>,<length>,<roffset>,<rlimit>) are optional. But these optional parameters must be entered in the specified order. For example, if you want to specify <length>, then you must also specify <soffset>. See details below for a definition of each of these parameters.

This command uses the data in the format specified by FORMat:DATA, returning either binary or ASCII data.

Example: To query the mean power of a set of GSM bursts:

Supply a signal that is a set of GSM bursts.

Select the IQ Waveform measurement (in IQ Analyzer Mode).

Set the sweep time to acquire at least one burst.

Set the triggers such that acquisition happens at a known position relative to a burst.

Then query the mean burst levels using, CALC:DATA2:COMP?

MEAN,24e–6,526e–6 (These parameter values correspond to GSM signals, where 526e–6 is the length of the burst in the slot and you just want 1 burst.)

Prior to A.02.00

Initial S/W Revision:

• BLOCk or block data - returns all the data points from the region of the trace data that you specify.

For example, it could be used to return the data points of an input signal over several timeslots, excluding the portions of the trace data that you do not want. (This is x,y pairs for trace data and I,Q pairs for complex data.)

• CFIT or curve fit - applies curve fitting routines to the data. <soffset> and <length> are required to define the data that you want. <roffset> is an optional parameter for the desired order of the curve equation. The query will return the following values: the x-offset (in seconds) and the curve coefficients ((order + 1) values).

MIN, MAX, MEAN, DME, RMS, RMSC, SAMP, SDEV and PPH return one data value for each specified region (or <length>) of trace data, for as many regions as possible until you run out of trace data (using <roffset> to specify regions). Or they return the number of regions you specify (using

<rlimit>) ignoring any data beyond that.

• MINimum - returns the minimum data point (x,y pair) for the specified region(s) of trace data. For

I/Q trace data, the minimum magnitude of the I/Q pairs is returned.

• MAXimum - returns the maximum data point (x,y pair) for the specified region(s) of trace data. For

I/Q trace data, the maximum magnitude of the I/Q pairs is returned.

160

Common Measurement Functions 1

NOTE

MEAN - returns a single value that is the arithmetic mean of the data point values

(in dB/ dBm) for the specified region(s) of trace data. For I/Q trace data, the mean of the magnitudes of the I/Q pairs is returned. See the following equations.

If the original trace data is in dB, this function returns the arithmetic mean of those log values, not log of the mean power which is a more useful value. The mean of the log is the better measurement technique when measuring CW signals in the presence of noise. The mean of the power, expressed in dB, is useful in power measurements such as Channel Power. To achieve the mean of the power, use the

RMS option.

Equation 1

Mean Value of Data Points for Specified Region(s)

MEAN =

_ n

Σ

Xi

Xi

∈ region(s)

vsd27-1 where Xi is a data point value, and n is the number of data points in the specified region(s).

Equation 2

Mean Value of I/Q Data Pairs for Specified Region(s)

MEAN =

_ n

Σ

|Xi|

Xi

∈ region(s) vsd27-2 where |Xi| is the magnitude of an I/Q pair, and n is the number of I/Q pairs in the specified region(s).

• DMEan - returns a single value that is the mean power (in dB/ dBm) of the data point values for the specified region(s) of trace data. See the following equation:

Equation 3

DMEan Value of Data Points for Specified Region(s)

DME = 10 x log

10 n

Σ

10

Xi

10

Xi

∈ region(s) vsd27-3

• RMS - returns a single value that is the average power on a root-mean-squared voltage scale

(arithmetic rms) of the data point values for the specified region(s) of trace data. See the following equation.

NOTE

For I/Q trace data, the rms of the magnitudes of the I/Q pairs is returned. See the following equation.

This function is very useful for I/Q trace data. However, if the original trace data

161

Common Measurement Functions 1

is in dB, this function returns the rms of the log values which is not usually needed.

Equation 4

RMS Value of Data Points for Specified Region(s)

RMS = n

Σ

Xi

2

Xi

∈ region(s) vsd27-4 where Xi is a data point value, and n is the number of data points in the specified region(s).

Equation 5

RMS Value of I/Q Data Pairs for Specified Region(s)

RMS = n

Σ

Xi Xi*

Xi

∈ region(s) vsd27-5 where Xi is the complex value representation of an I/Q pair, Xi* its conjugate complex number, and n is the number of I/Q pairs in the specified region(s).

Once you have the rms value for a region of trace data (linear or I/Q), you may want to calculate the mean power. You must convert this rms value (peak volts) to power in dBm:

10 x log[10 x (rms value)2]

• SAMPle - returns the first data value (x,y pair) for the specified region(s) of trace data. For I/Q trace data, the first I/Q pair is returned.

• SDEViation - returns a single value that is the arithmetic standard deviation for the data point values for the specified region(s) of trace data. See the following equation.

For I/Q trace data, the standard deviation of the magnitudes of the I/Q pairs is returned. See the following equation.

Equation 6

Standard Deviation of Data Point Values for Specified Region(s)

SDEV = n

Σ

(Xi – X)

2

Xi

∈ region(s) vsd27-7 where Xi is a data point value, X is the arithmetic mean of the data point values for the specified region(s), and n is the number of data points in the specified region(s).

SDEV = n

Σ

(|Xi| – X)

Xi

∈ region(s)

2 vsd27-8 where |Xi| is the magnitude of an I/Q pair, X is the mean of the magnitudes for the specified region(s),

162

Common Measurement Functions 1

and n is the number of data points in the specified region(s).

• PPHase - returns the x,y pairs of both rms power (dBm) and arithmetic mean phase (radian) for every specified region and frequency offset (Hz). The number of pairs is defined by the specified number of regions. This parameter can be used for I/Q vector (n=0) in Waveform (time domain) measurement and all parameters are specified by data point in PPHase.

The rms power of the specified region may be expressed as:

Power = 10 x log [10 x (RMS I/Q value)] + 10.

The RMS I/Q value (peak volts) is: n

Σ

Xi

Xi Xi*

∈ region

vsd27-9 where Xi is the complex value representation of an I/Q pair, Xi* its conjugate complex number, and n is the number of I/Q pairs in the specified region.

The arithmetic mean phase of the specified region may be expressed as: n

Σ

Yi

Yi

∈ region

vsd27-10 where Yi is the unwrapped phase of I/Q pair with applying frequency correction and n is the number of I/Q pairs in the specified region.

The frequency correction is made by the frequency offset calculated by the arithmetic mean of every specified region’s frequency offset. Each frequency offset is calculated by the least square method against the unwrapped phase of I/Q pair.

Sample Trace Data - Constant Envelope

(See below for explanation of variables.) length t

0 soffset roffset

Sample Trace Data - Not Constant Envelope

If rlimit is set to 3, this last chunk of data will be ignored.

vsd29

163

Common Measurement Functions 1

(See below for explanation of variables.) length

If rlimit is set to 3, this chunk of data and any additional data will be ignored.

soffset t

0 roffset vsd30

<soffset> - start offset is an optional real number. (It is in seconds for time-domain traces, and is a dimensionless index 0 to Npoints – 1, for frequency-domain traces). It specifies the amount of data at the beginning of the trace that will be ignored before the decimation process starts. It is the time or frequency change from the start of the trace to the point where you want to start using the data. The default value is zero.

<length> - is an optional real number. (It is in seconds for time-domain traces, and is a dimensionless index 0 to Npoints – 1, for frequency-domain traces). It defines how much data will be compressed into one value. This parameter has a default value equal to the current trace length.

<roffset> - repeat offset is an optional real number. (It is in seconds for time-domain traces, and is a dimensionless index 0 to Npoints – 1, for frequency-domain traces). It defines the beginning of the next field of trace elements to be compressed. This is relative to the beginning of the previous field.

This parameter has a default value equal to the <length> variable. Note that this parameter is used for a completely different purpose when curve fitting (see CFIT above).

<rlimit> - repeat limit is an optional integer. It specifies the number of data items that you want returned. It will ignore any additional items beyond that number. You can use the Start offset and the

Repeat limit to pick out exactly what part of the data you want to use. The default value is all the data.

Calculate Peaks of Trace Data (Remote Command Only)

Returns a list of all the peaks for the currently selected measurement and sub-opcode [n]. The peaks must meet the requirements of the peak threshold and excursion values.

n = any valid sub-opcode for the current measurement. See the MEASure:<measurement> command description of your specific measurement for information on the data that can be returned.

The command can only be used with specific sub-opcodes with measurement results that are trace data.

Both real and complex traces can be searched, but complex traces are converted to magnitude in dBm. In many measurements the sub-opcode n=0, is the raw trace data which cannot be searched for peaks. And

Sub-opcode n=1, is often calculated results values which also cannot be searched for peaks.

This command uses the data setting specified by the FORMat:BORDer and FORMat:DATA commands and can return real or ASCII data. If the format is set to INT,32, it returns REAL,32 data.

The command has four types of parameters:

164

Common Measurement Functions 1

• Threshold (in dBm)

• Excursion (in dB)

• Sorting order (amplitude, frequency, time)

• Optional in some measurements: Display line use (all, > display line, < display line)

Remote Command:

Example:

For Swept SA measurement:

:CALCulate:DATA[1]|2|3|4|5|6:PEAKs?

<threshold>,<excursion>[,AMPLitude|FREQuency|TIME[,ALL|

GTDLine|LTDLine]]

For most other measurements:

:CALCulate:DATA[1]|2|3|4|5|6:PEAKs?

<threshold>,<excursion>[,AMPLitude|FREQuency|TIME]

Example for Swept SA measurement in Spectrum Analyzer Mode:

CALC:DATA4:PEAK? –40,10,FREQ,GTDL This will identify the peaks of trace 4 that are above –40 dBm, with excursions of at least 10 dB. The peaks are returned in order of increasing frequency, starting with the lowest frequency. Only the peaks that are above the display line are returned.

Query Results 1:

With FORMat:DATA REAL,32 selected, it returns a list of floating-point numbers. The first value in the list is the number of peak points that are in the following list. A peak point consists of two values: a peak amplitude followed by its corresponding frequency (or time).

If no peaks are found the peak list will consist of only the number of peaks,

(0).

165

Common Measurement Functions 1

Notes: <n> - is the trace that will be used

<threshold> - is the level below which trace data peaks are ignored. Note that the threshold value is required and is always used as a peak criterion. To effectively disable the threshold criterion for this command, provide a substantially low threshold value such as –200 dBm. Also note that the threshold value used in this command is independent of and has no effect on the threshold value stored under the Peak Criteria menu.

<excursion> - is the minimum amplitude variation (rise and fall) required for a signal to be identified as peak. Note that the excursion value is required and is always used as a peak criterion. To effectively disable the excursion criterion for this command, provide the minimum value of 0.0 dB. Also note that the excursion value used in this command is independent of and has no effect on the excursion value stored under the Peak Criteria menu.

Values must be provided for threshold and excursion. The sorting and display line parameters are optional (defaults are AMPLitude and ALL).

Note that there is always a Y-axis value for the display line, regardless of whether the display line state is on or off. It is the current Y-axis value of the display line which is used by this command to determine whether a peak should be reportedSorting order:

AMPLitude - lists the peaks in order of descending amplitude, with the highest peak first (default if optional parameter not sent)

FREQuency - lists the peaks in order of occurrence, left to right across the x-axis.

TIME - lists the peaks in order of occurrence, left to right across the x-axis.

Peaks vs. Display Line:

ALL - lists all of the peaks found (default if optional parameter not sent).

GTDLine (greater than display line) - lists all of the peaks found above the display line.

LTDLine (less than display line) - lists all of the peaks found below the display line.

Prior to A.02.00

Initial S/W Revision:

Format Data: Numeric Data (Remote Command Only)

This command specifies the format of the trace data input and output. It specifies the formats used for trace data during data transfer across any remote port. It affects only the data format for setting and querying trace data for the :TRACe[:DATA], TRACe[:DATA]?, :CALCulate:DATA[n]? and

FETCh:SANalyzer[n]? commands and queries.

Remote Command:

:FORMat[:TRACe][:DATA] ASCii|INTeger,32|REAL,32 |REAL,64

:FORMat[:TRACe][:DATA]?

166

Notes:

Dependencies:

Common Measurement Functions 1

The query response is:

ASCii: ASC,8

REAL,32: REAL,32

REAL,64: REAL,64

INTeger,32: INT,32

When the numeric data format is REAL or ASCii, data is output in the current

Y Axis unit. When the data format is INTeger, data is output in units of m dBm (.001 dBm).

Note that the INT,32 format is only applicable to the command,

TRACe:DATA. This preserves backwards compatibility for the Swept SA measurement. For all other commands/queries which honor FORMat:DATA, if INT,32 is sent the analyzer will behave as though it were set to REAL,32.

The INT,32 format returns binary 32-bit integer values in internal units (m dBm), in a definite length block.

Sending a data format spec with an invalid number (for example, INT,48) generates no error. The analyzer simply uses the default (8 for ASCii, 32 for

INTeger, 32 for REAL).

Sending data to the analyzer which does not conform to the current FORMat specified, results in an error.

Preset:

Initial S/W Revision:

ASCii

Prior to A.02.00

The specs for each output type follow:

ASCii - Amplitude values are in ASCII, in the current Y Axis Unit, one ASCII character per digit, values separated by commas, each value in the form:

SX.YYYYYEsZZ

Where:

S = sign (+ or -)

X = one digit to left of decimal point

Y = 5 digits to right of decimal point

E = E, exponent header s = sign of exponent (+ or -)

ZZ = two digit exponent

REAL,32 - Binary 32-bit real values in the current Y Axis Unit, in a definite length block.

REAL,64 - Binary 64-bit real values in the current Y Axis Unit, in a definite length block.

167

Common Measurement Functions 1

Format Data: Byte Order (Remote Command Only)

This command selects the binary data byte order for data transfer and other queries. It controls whether binary data is transferred in normal or swapped mode. This command affects only the byte order for setting and querying trace data for the :TRACe[:DATA], TRACe[:DATA]? , :CALCulate:DATA[n]? and

FETCh:SANalyzer[n]? commands and queries.

By definition any command that says it uses FORMat:DATA uses any format supported by

FORMat:DATA.

The NORMal order is a byte sequence that begins with the most significant byte (MSB) first, and ends with the least significant byte (LSB) last in the sequence: 1|2|3|4. SWAPped order is when the byte sequence begins with the LSB first, and ends with the MSB last in the sequence: 4|3|2|1.

Remote Command:

Preset:

Initial S/W Revision:

:FORMat:BORDer NORMal|SWAPped

:FORMat:BORDer?

NORMal

Prior to A.02.00

Meas Setup

The Meas Setup key opens up a menu of softkeys that allow you to control the most important parameters for the current measurement.

NOTE

In the Meas Setup menu you may configure Averaging, by setting the Average

Number and the Average Type.

Key Path

Initial S/W Revision

Front-panel key

Prior to A.02.00

Average/Hold Number

Sets the terminal count number N for

Average, Max Hold

and

Min Hold

trace types. This number is an integral part of how the average trace is calculated. Basically, increasing N results in a smoother average trace.

See

“More Information” on page 169 .

See

“AVER:CLE command” on page 169 .

Key Path

Remote Command

Meas Setup

[:SENSe]:AVERage:COUNt <integer>

[:SENSe]:AVERage:COUNt?

168

Common Measurement Functions 1

Couplings

Preset

State Saved

Min

Max

Status Bits/OPC dependencies

Initial S/W Revision

Restarting any of these functions (

Average, Max Hold

or

Min Hold)

restarts all of them, as there is only one count.

100

Saved in Instrument State

1

10000

See the Section

“Sweep/Control” on page 263 for a discussion of the

Sweeping, Measuring, Settling and OPC bits, and the Hi Sweep line. All are affected when a sequence is reset.

Prior to A.02.00

More Information

When in

Single

, the sweep stops when N is reached. If you want to add one more Average, or one more trace to Max Hold or Min Hold, simply press the Up key while Average/Hold Number is the active function.

In

Cont

(continuous), averaging and holding continues even after N is reached. Therefore, using doing trace holding in

Cont

, the value of N is irrelevant. But for averaging, each new sweep is exponentially averaged in with a weighting equal to N.

For details of how the average trace is calculated and how this depends on the

Average/Hold Number

,

see “Average Type” on page 170

. For details on how the various control functions in the instrument start and restart averaging. See the Section

“Sweep/Control” on page 263 for a discussion of the Sweeping,

Measuring, Settling and OPC bits, and the Hi Sweep line. All are affected when a sequence is reset.

The Average/Hold Number is not affected by Auto Couple.

AVER:CLE command

The AVER:CLE command (below) resets the average/hold count and does an INIT:IMM, which begins another set of sweeps when trigger conditions are satisfied. It only does this if an active trace is in

Average or Hold type.

Remote Command:

Example:

Notes:

Default Unit:

Initial S/W Revision:

[:SENSe]:AVERage:CLEar

AVER:COUN 100

AVER:CLE sets the current count (k and K) to 1 and restarts the averaging process.

When the instrument receives this command it performs an INIT:IMM, if and only if there is an active trace in Max Hold, Min Hold, or Average type.

Enter

Prior to A.02.00

169

Common Measurement Functions 1

Average Type

Lets you control the way averaging is done by choosing one of the following averaging scales: log-power (video), power (RMS), or voltage averaging. Also lets you choose Auto Average Type

(default).

When performing Trace Averaging, , the equation that is used to calculate the averaged trace depends on the average type. See the descriptions for the keys which select each Average Type (

“Log-Pwr Avg

(Video)” on page 171

(Video), “Pwr Avg (RMS)” on page 172

(RMS), or

“Voltage Avg” on page 172

) for details on these equations.

See

“More Information” on page 170 .

Key Path

Remote Command

Preset

State Saved

Readback line

Initial S/W Revision

Meas Setup

[:SENSe]:AVERage:TYPE:AUTO OFF|ON|0|1

[:SENSe]:AVERage:TYPE:AUTO?

ON

Saved in Instrument State

1-of-N selection as

Log-Pwr (Video) for Log-Pwr (Video) Avg

Pwr (RMS) for Power Avg

Voltagefor Voltage

Prior to A.02.00

Remote Command:

Notes:

[:SENSe]:AVERage:TYPE RMS|LOG|SCALar

[:SENSe]:AVERage:TYPE?

Parameters map to avg types as:

RMS = Pwr (RMS) Avg

LOG = Log-Pwr (Video) Avg

SCALar = Voltage Avg

LOG Preset:

Backwards Compatibility SCPI: [:SENSe]:AVERage:TYPE LINear sets Scalar averaging

[:SENSe]:AVERage:TYPE VOLTage sets Scalar averaging

[:SENSe]:AVERage:TYPE VIDeo sets Log-Power averaging

[:SENSe]:AVERage:TYPE LPOWer sets Log-Power averaging

[:SENSe]:AVERage:TYPE POWer sets RMS averaging

Initial S/W Revision: Prior to A.02.00

More Information

When you select log-power averaging, the measurement results are the average of the signal level in

170

Common Measurement Functions 1

logarithmic units (decibels). When you select power average (RMS), all measured results are converted into power units before averaging and filtering operations, and converted back to decibels for displaying.

Remember: there can be significant differences between the average of the log of power and the log of the average power.

These are the averaging processes within a spectrum analyzer and all of them are affected by this setting:

Trace averaging (see Section

“Trace/Detector” on page 428

”) averages signal amplitudes on a trace-to-trace basis. The average type applies to all traces in Trace Average (it is not set on a trace-by-trace basis).

Average detector (see Section

“Trace/Detector” on page 428

”) averages signal amplitudes during the time or frequency interval represented by a particular measurement point.

Noise Marker (see Section “Marker Function” on page 1593 ”) averages signal amplitudes across measurement points to reduce variations for noisy signals.

VBW filtering (see Section “BW” on page 1473 ”) adds video filtering which is a form of averaging of the video signal.

When

Auto

is selected, the analyzer chooses the type of averaging (see below). When one of the average types is selected manually, the analyzer uses that type regardless of other analyzer settings, and shows

Man on the

Average Type

softkey.

Auto

Chooses the optimum type of averaging for the current instrument measurement settings.

Key Path

Example

Notes

Couplings

State Saved

Readback

Meas setup, Average Type

AVER:TYPE:AUTO ON

See

Average Type

, above

Here are the auto-select rules for

Average Type:

Auto

selects

Voltage Averaging

if the Detector for any active trace is

EMI

Average

or

QPD or RMS Average

; otherwise it selects

Power (RMS)

Averaging if a

Marker Function

(

Marker Noise

,

Band/Intvl Power

) is on, or

Detector

is set to

Man

and

Average

; otherwise if

Amplitude

,

Scale Type

is set to

Lin

it selects

Voltage

Averaging; otherwise, if the EMC Standard is set to CISPR, it selects Voltage; otherwise

Auto

selects

Log-Power

Average.

Note that these rules are only applied to active traces. Traces which are not updating do not impact the auto-selection of Average Type.

Saved in Instrument State

The type auto-selected is displayed in the readback line on the

Average Type

key

Prior to A.02.00

Initial S/W Revision

Log-Pwr Avg (Video)

Selects the logarithmic (decibel) scale for all filtering and averaging processes. This scale is sometimes called “Video” because it is the most common display and analysis scale for the video signal within a

171

Common Measurement Functions 1

spectrum analyzer. This scale is excellent for finding CW signals near noise, but its response to noise-like signals is 2.506 dB lower than the average power of those noise signals. This is compensated for in the Marker Noise function.

The equation for trace averaging on the log-pwr scale is shown below, where K is the number of averages accumulated. (In continuous sweep mode, once K has reached the Average/Hold Number, K stays at that value, providing a continuous running average.)

New avg = ((K–1)Old avg + New data)/K

Assumes all values in decibel scale.

Key Path

Example

Notes

Couplings

Readback

Initial S/W Revision

Meas setup, Average Type

AVER:TYPE LOG

See “ “Average Type” on page 170 ”

See “ “Auto” on page 171

Log-Pwr (Video)

Prior to A.02.00

Pwr Avg (RMS)

In this average type, all filtering and averaging processes work on the power (the square of the magnitude) of the signal, instead of its log or envelope voltage. This scale is best for measuring the true time average power of complex signals. This scale is sometimes called RMS because the resulting voltage is proportional to the square root of the mean of the square of the voltage.

In the equation for averaging on this scale (below), K is the number of averages accumulated. (In continuous sweep mode, once K has reached the Average/Hold Number, K stays at that value, providing a running average.)

New avg = 10 log ((1/K)((K–1)(10Old avg/10)+10New data/10))

Equation assumes all values are in the decibel scale.

Key Path

Example

Notes

Couplings

Readback

Initial S/W Revision

Meas setup, Average Type

AVER:TYPE RMS

See “Average Type” on page 170

See “Auto” on page 171

Pwr (RMS)

Prior to A.02.00

Voltage Avg

In this Average type, all filtering and averaging processes work on the voltage of the envelope of the signal. This scale is good for observing rise and fall behavior of AM or pulse-modulated signals such as radar and TDMA transmitters, but its response to noise-like signals is 1.049 dB lower than the average power of those noise signals. This is compensated for in the

Marker Noise

function.

172

Common Measurement Functions 1

In the equation for averaging on this scale (below), K is the number of averages accumulated. (In continuous sweep mode, once K has reached the Average/Hold Number, K stays at that value.)

New avg = 20 log ((1/K)((K–1)(10Old avg/20)+10New data/20))

Equation assumes all values are in the decibel scale.

Key Path

Example

Notes

Couplings

Readback

Initial S/W Revision

Meas setup, Average Type

AVER:TYPE SCAL

See

“Average Type” on page 170

See

“Auto” on page 171

Pwr (RMS)

Prior to A.02.00

Limits

The limits softkey opens up a menu of softkeys to control the limits for the current measurement. Limits arrays can be entered by the user, sent over SCPI, or loaded from a file.

Initial S/W Revision

Key Path

Dependencies

Preset

A.02.00

Meas Setup

This key will only appear if you have the proper option installed in your instrument.

Limits are turned off by a Preset, but the Limits arrays (data) are only reset

(deleted) by Restore Mode Defaults. They survive shutdown and restarting of the analyzer application, which means they will survive a power cycle.

Select Limit

Specifies the selected limit. The term “selected limit” is used throughout this document to specify which limit will be affected by the functions.

Key Path

Notes

Preset

State Saved

Initial S/W Revision

Meas Setup, Limits

The selected limit is remembered even when not in the Limit Menu.

Limit 1, not affected by Mode Preset, preset by Restore Mode Defaults.

Saved in State.

A.02.00

Limit On/Off

Selects whether the limit and margin are displayed. If Test Limits is on, this also determines whether the

test trace (see “Test Trace” on page 175 ) will be tested against the limit. If

Limit On/Off

is

On

, the following occurs:

173

Common Measurement Functions 1

• The limit line is displayed, in the same color as the limited trace, but paler. Portions of traces which fail the limits will be displayed in red.

• The margin line is displayed if Margin is on and the Margin Value is non-zero (see

“Margin” on page 180

). The margin line is displayed in the same color as the limit line, but paler still and dashed.

Portions of traces which pass the limits but fail the margin will be displayed in amber.

• The trace is tested for the purpose of the “Trace Pass/Fail”indication in the graticule if, in addition to

Limit On/Off

being

On,

the trace is displayed and

Test Limits (All Limits)

is on (see “Test Limits” on page 186

). If the trace is not tested, no report of the trace passing or failing is seen on the graticule.

Note that the SCPI queries of Limit Pass/Fail are independent of these conditions; the test is always performed when queried over SCPI.

The PASS/FAIL box in the corner of the Meas Bar is only displayed if there is at least one “Trace

Pass/Fail” indication displayed in the graticule.

Note that the red and amber coloring of traces which fail the limits and/or margins only applies to traces whose X-axis corresponds to the current analyzer X-axis. Traces which are not updating (in View, for example) will not change color if the analyzer X-axis settings (eg, start and stop frequency) do not match those of the trace, for example if they have been changed since the trace stopped updating. In this case, the Invalid Data indicator (*) will appear in the upper right hand corner.

When the limits are frequency limits but the trace is a zero-span trace, the limit trace is drawn at the limit amplitude of the center frequency. When the limits are time limits but the trace is a frequency domain trace, the limit trace is drawn according to the current time axis, with the left of the screen being 0 and the right being equal to sweep time.

Key Path

Remote Command

Example

Dependencies

Couplings

Preset

State Saved

Backwards Compatibility SCPI

Initial S/W Revision

Meas Setup, Limits

:CALCulate:LLINe[1]|2|3|4|5|6:DISPlay OFF|ON|0|1

:CALCulate:LLINe[1]|2|3|4|5|6:DISPlay?

:CALC:LLIN2:DISP ON !turns on the display for limit line 2.

This command will generate an “Option not available” error unless you have the proper option installed in your instrument.

Limit display ON selects the limit.

Testing is done on all displayed limits if

Test Limits (All Limits)

is ON.

Entering the limit menu from the GUI turns on the selected limit.

OFF

Saved in State.

:CALCulate:LLINe[1]|2:STATe OFF|ON|0|1

(In the past you had to send the DISP command as well as the STATe command in order to get a limit on and testing. Now, the DISP command is sufficient, but we accept the state command and do nothing with it)

A.02.00

174

Common Measurement Functions 1

Properties

Accesses a menu which lets you set the properties of the selected limit.

Key Path

Initial S/W Revision

Meas Setup, Limits

A.02.00

Select Limit

Specifies the selected limit. The term “selected limit” is used throughout this document to specify which limit will be affected by the functions.

Key Path

Notes

Preset

State Saved

Initial S/W Revision

Meas Setup, Limits, Properties

The selected limit is remembered even when not in the Limit Menu.

Limit 1, not affected by Mode Preset, preset by Restore Mode Defaults.

Saved in State.

A.02.00

Test Trace

Selects the trace you want the limit to test. A limit is applied to one and only one trace; each trace can have both an upper and a lower limit. When executing Limit Test, the limit is applied only to the specified trace.

A trace can have multiple limit lines simultaneously; in that case, only one upper and one lower limit line will affect the color of the trace. Other limit lines will be displayed, and will affect the pass/fail status, but the trace will not turn red if it crosses a secondary limit line.

Key Path

Remote Command

Example

Notes

Couplings

Preset

State Saved

Min

Max

Readback

Meas Setup, Limits, Properties

:CALCulate:LLINe[1]|2|3|4|5|6:TRACe 1|2|3|4|5|6

:CALCulate:LLINe[1]|2|3|4|5|6:TRACe?

:CALC:LLIN3:TRAC 2 applies limit 3 to trace 2.

When the trace display is off, the trace is not tested. The trace is tested only when the trace display is on and Test Limits (see

“Test Limits” on page 186

) is on.

This matters when testing a trace or limit line for failure, via

:CALC:LLIN3:FAIL? or :CALC:TRAC2:FAIL?

Limits 1 and 2 preset to 1, Limits 3 and 4 preset to 2, Limits 5 and 6 preset to 3

Not affected by Mode Preset, preset by Restore Mode Defaults.

Saved in State.

1

6

Trace 1|2|3|4|5|6

175

Common Measurement Functions 1

Initial S/W Revision A.02.00

Type

Selects whether the limit you are editing is an upper or lower limit. An upper limit fails if the trace exceeds the limit. A lower limit fails if the trace falls below the limit.

Key Path

Remote Command

Example

Couplings

Preset

State Saved

Initial S/W Revision

Meas Setup, Limits, Properties

:CALCulate:LLINe[1]|2|3|4|5|6:TYPE UPPer|LOWer

:CALCulate:LLINe[1]|2|3|4|5|6:TYPE?

:CALC:LLIN2:TYPE LOW !sets limit line 2 to act as a lower limit.

If a margin has already been set for this limit line, and this key is used to change the limit type, then the margin value will reverse sign.

Upper for Line 1, 3, and 5; Lower for Line 2, 4, 6.

Not affected by Mode Preset, preset by Restore Mode Defaults.

Saved in State.

A.02.00

Interpolation

Accesses a menu which lets you set the frequency and amplitude interpolation of the selected limit.

Key Path

Readback

Initial S/W Revision

Meas Setup, Limits, Properties

[Lin|Log Frequency, Lin|Log Amplitude]

A.02.00

Frequency Interpolation

This key is grayed out if Time is the selected X Axis Units. Sets the interpolation between frequency points, allowing you to determine how limit trace values are computed between points in a limit table. The available interpolation modes are linear and logarithmic. If frequency interpolation is logarithmic (Log), frequency values between limit points are computed by first taking the logarithm of both the table values and the intermediate value.

A linear interpolation is then performed in this logarithmic frequency space. An exactly analogous manipulation is done for logarithmic amplitude interpolation.

Note that the native representation of amplitude is in dB.

For linear amplitude interpolation and linear frequency interpolation, the interpolation is computed as:

y

=

y i

+

1

20 log(

10 20

f i

+

1

y i

10

f i

20

(

f

f i

)

y i

+

10

20

)

For linear amplitude interpolation and log frequency interpolation, the interpolation is computed as:

176

Common Measurement Functions 1

y

=

y i

+

1

20

10 log( log

20

f i

+

1

y i

10

20 log

f i

( log

f

− log

f i

)

y i

+

10 20 )

For log amplitude interpolation and linear frequency interpolation, the interpolation is computed as:

y

=

y i

+

1

f i

+

1

y i f i

(

f

f i

)

+

y i

For log amplitude interpolation and log frequency interpolation, the interpolation is computed as:

y

= log

y i

+

1

f i

+

1

y i

log

f i

(

log

f

− log

f i

)

+

y i

NOTE

Interpolation modes determine how limit values are computed between points in the limit table. The appearance of a limit trace is also affected by the amplitude scale, which may be linear or logarithmic.

Key Path

Remote Command

Example

Preset

State Saved

Initial S/W Revision

Meas Setup, Limits, Properties, Interpolation

:CALCulate:LLINe[1]|2|3|4|5|6:CONTrol:INTerpolate:TYPE

LOGarithmic|LINear

:CALCulate:LLINe[1]|2|3|4|5|6:CONTrol:INTerpolate:TYPE?

:CALC:LLIN:CONT:INT:TYPE LIN !sets limit line 1 frequency interpolation to linear.

Linear, not affected by Mode Preset, preset by Restore Mode Defaults.

Saved in instrument state.

A.02.00

Amplitude Interpolation

Sets the interpolation to linear or logarithmic for the specified limiting points set, allowing you to determine how limit trace values are computed between points in a limit table. See Frequency Interpolation for the equations used to calculate limit values between points.

Key Path

Remote Command

Example

Preset

Meas Setup, Limits, Properties, Interpolation

:CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:INTerpolate:TYP

E LOGarithmic|LINear

:CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:INTerpolate:TYP

E?

:CALC:LLIN:AMPL:INT:TYPE LIN !sets limit line 1 amplitude interpolation to linear.

Logarithmic, not affected by Mode Preset, preset by Restore Mode Defaults.

177

Common Measurement Functions 1

State Saved

Initial S/W Revision

Saved in instrument state.

A.02.00

Fixed / Relative

Opens a menu which will allow you to specify that the selected limit is relative to either Center Frequency or

Reference level.

Key Path

Readback

Initial S/W Revision

Meas Setup, Limits, Properties

Fixed|Rel to CF|Rel to RL|Rel to CF + RL (square brackets)

A.02.00

Relative to CF

Chooses whether the limit line frequency points are coupled to the instrument center frequency, and whether the frequency points are expressed as an offset from the instrument center frequency. If the limit lines are specified with time, this has no effect. The limit table must in this case support negative frequencies.

For example, assume you have a frequency limit line, and the analyzer center frequency is at 1 GHz. If Relative to

CF is “Off”, entering a limit line segment with a frequency coordinate of 300 MHz displays the limit line segment at 300 MHz, and the limit line segment will not change frequency if the center frequency changes. If Relative to CF is “On”, entering a limit line segment with a frequency coordinate of 300 MHz displays the limit line segment at

CF + 300 MHz, or 1.3 GHz. Furthermore, if the center frequency changes to 2 GHz, the limit line segment will be displayed at CF + 300 MHz, or 2.3 GHz.

It is possible to change this setting after a limit line has been entered. When changing from On to Off or vice-versa, the frequency values in the limit line table change so that the limit line remains in the same position for the current frequency settings of the analyzer.

Pressing this button makes Center Frequency the active function.

Key Path

Remote Command

Example

Notes

Couplings

Preset

State Saved

Initial S/W Revision

Meas Setup, Limits, Properties, Fixed/Relative

:CALCulate:LLINe[1]|2|3|4|5|6:FREQuency:CMODe:RELative

ON|OFF|1|0

:CALCulate:LLINe[1]|2|3|4|5|6:FREQuency:CMODe:RELative?

:CALC:LLIN:FREQ:CMOD:REL ON !makes limit line 1 relative to the center frequency.

If the Trace Domain is changed to Time

(:CALCulate:LLINe:CONTrol:DOMain TIME), the SCPI command :

:CALCulate:LLINe[1]|2|3|4|5|6:FREQuency:CMODe:RELative ON|OFF|1|0 will have no effect.

Pressing this button makes Center Frequency the active function.

Off, not affected by Mode Preset, preset by Restore Mode Defaults.

Saved in instrument state.

A.02.00

178

Common Measurement Functions 1

Relative to RL

Chooses whether the limit line amplitude points are coupled to the instrument reference level, and whether the amplitude points are expressed as an offset from the instrument reference level.

For example, assume you have a limit line, and the reference level at –10 dBm. If Relative to RL is “Off”, entering a limit line segment with an amplitude coordinate of –20 dB displays the limit line segment at –20 dBm, and the limit line segment will not change amplitude if the reference level amplitude changes. If Relative to RL is “On”, entering a limit line segment with an amplitude coordinate of –20 dB displays the limit line segment at RL – 20 dB, or –30 dBm. Furthermore, if the reference level amplitude changes to –30 dBm, the limit line segment will be displayed at RL – 20 dB, or –50 dBm.

It is possible to change this setting after a limit line has been entered. When changing from On to Off or vice-versa, the amplitude values in the limit line table change so that the limit line remains in the same position for the current reference level settings of the analyzer.

Key Path

Remote Command

Example

Couplings

Preset

State Saved

Initial S/W Revision

Meas Setup, Limits, Properties, Fixed/Relative

:CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:CMODe:RELative

ON|OFF|1|0

:CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:CMODe:RELative?

:CALC:LLIN:AMPL:CMOD:REL ON !makes limit line 1 relative to the reference level amplitude.

Pressing this button makes Reference level the active function.

Off, not affected by Mode Preset, preset by Restore Mode Defaults.

Saved in instrument state.

A.02.00

Description

Provides a description of up to 60 characters by which the operator can easily identify the limit. Will be stored in the exported file. Can be displayed in the active function area by selecting as the active function, if desired to be in a screen dump.

Key Path

Remote Command

Example

Preset

State Saved

Readback

Initial S/W Revision

Meas Setup, Limits, Properties

:CALCulate:LLINe[1]|2|3|4|5|6:DESCription “Description”

:CALCulate:LLINe[1]|2|3|4|5|6:DESCription?

:CALC:LLIN:DESC “European Emissions”

“” (null String), not affected by Mode Preset, preset by Restore Mode

Defaults.

Saved in instrument stat.e

As much of the description will fit on one line of the key, followed by “…” if some of the description will not fit on one line of the key.

A.02.00

179

Common Measurement Functions 1

Comment

Sets an ASCII comment field which will be stored in an exported file. Can be displayed in the active function area by selecting as the active function, if desired to be in a screen dump. The Limits .csv file supports this field.

Key Path

Remote Command

Example

Preset

State Saved

Readback

Initial S/W Revision

Meas Setup, Limits, Properties

:CALCulate:LLINe[1]|2|3|4|5|6:COMMent "text"

:CALCulate:LLINe[1]|2|3|4|5|6:COMMent?

:CALC:LLIN1:COMM "this is a comment"

“” (null String), not affected by Mode Preset, preset by Restore Mode

Defaults.

Saved in instrument state.

As much of the description will fit on one line of the key, followed by “…” if some of the description will not fit on one line of the key.

A.02.00

Margin

Selects a margin for this limit, which will cause a trace to Fail Margin when the trace is between the limit line and the margin line. Portions of the traces which pass the limit but fail the margin will be displayed in an amber color.

A margin is always specified in dB relative to a limit – an upper limit will always have a negative margin, and a lower limit will always have a positive margin. If a value is entered with the incorrect sign, the system will automatically take the negative of the entered value.

If the limit type is switched from lower to upper while margin is present, the margin will reverse sign.

When the Margin is selected, it may be turned off by pressing the Margin key until Off is underlined.

This may also be done by performing a preset. Margin is the default active function whenever the margin is on, and it is not the active function whenever the margin is off.

The margin lines are displayed in the same color as limit lines, but paler. . If the limited trace is blanked then the limit line and the margin line will be blanked as well.

Key Path

Remote Command

Meas Setup, Limits

:CALCulate:LLINe[1]|2|3|4|5|6:MARGin <rel_ampl>

:CALCulate:LLINe[1]|2|3|4|5|6:MARGin?

:CALCulate:LLINe[1]|2|3|4|5|6:MARGin:STATe OFF|ON|0|1

:CALCulate:LLINe[1]|2|3|4|5|6:MARGin:STATe?

180

Common Measurement Functions 1

Example

Notes

Couplings

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

:CALC:LLIN1:MARG –2dB !sets limit line 1’s margin to –2 dB (Limit Line 1 is by default an upper limit).

:CALC:LLIN2:MARG 1dB !sets limit line 2’s margin to 1 dB (Limit Line 2 is by default a lower limit).

:CALC:LLIN2:MARG:STAT OFF !turns off the margin for limit line 2 and removes any tests associated with that margin line.

The queries “Limit Line Fail?” (:CALCulate:LLINe[1]|2|3|4|5|6:FAIL?) and

“Trace Fail?” (:CALCulate:TRACe[1]|2|3|4|5|6:FAIL?) will return 1 if the margin fails.

This will affect :CALC:LLIN3:FAIL or :CALC:TRAC2:FAIL?

not affected by Mode Preset, set to 0 dB for all Limits by Restore Mode

Defaults.

Saved in instrument state.

–40 dB (Upper); 0 dB (Lower)

0 dB (Upper); 40 dB (Lower); dB

A.02.00

Edit

Opens the Table Editor for the selected limit line.

When entering the menu, the editor window (with the limit table) turns on, the selected Limit is turned

On

and the amplitude scale is set to

Log.

The display of the trace to which the selected limit applies is turned on (thus, traces in Blank are set to View and traces in Background are set to On).

Turning on the

Limit means it’s display will be on, and it’s testing mode will be on as well; you should turn off any other limits that are on if they interfere with the editing of the selected limit.

NOTE

The table editor will only operate properly if the analyzer is sweeping, because its updates are tied to the sweep system. Thus, you should not try to use the editor in single sweep, and it will be sluggish during compute-intensive operations like narrow-span FFT sweeps.

When exiting the edit menu (by using the Return key or by pressing an instrument front panel key), the editor window turns off, however the Limit is still on and displayed, and the amplitude scale remains

Log

.

Limits are turned off by a Preset, but the Limits arrays (data) are only reset (deleted) by Restore Mode

Defaults. They survive shutdown and restarting of the analyzer application, which means they will survive a power cycle.

Key Path

Meas Setup, Limits

181

Common Measurement Functions 1

Couplings Turns the Limit Peaks table off.

A remote user can enter or access limit line data via

:CALCulate:LLINe[1]|2|3|4|5|6:DATA

A.02.00

Initial S/W Revision

Navigate

Lets you move through the table to edit the desired point

Key Path

Notes

Initial S/W Revision

Meas Setup, Limits, Edit

There is no value readback on the key

A.02.00

Frequency

Lets you edit the frequency of the current row.

Key Path

Notes

Initial S/W Revision

Meas Setup, Limits, Edit

There is no value readback on the key

A.02.00

Amplitude

Lets you edit the Amplitude of the current row.

Key Path

Notes

Min

Max

Initial S/W Revision

Meas Setup, Limits, Edit

There is no value readback on the key

–1000 dBm

1000 dBm

A.02.00

Insert Point Below

Pressing this key inserts a point below the current point. The new point is a copy of the current point. And becomes the current point The new point is not yet entered into the underlying table, and the data in the row is displayed in

LightGray.

Key Path

Initial S/W Revision

Meas Setup, Limits, Edit

A.02.00

Delete Point

This is an immediate action key. It will immediately delete the currently-selected point, whether or not that point is being edited, and select Navigate. The point following the currently-selected point (or the point preceeding if there

182

Common Measurement Functions 1

is none) will be selected.

Key Path

Initial S/W Revision

Meas Setup, Limits, Edit

A.02.00

Copy from Limit

Copies an existing limit into the current limit, including all secondary parameters (Description, Associated Trace,

Type, Margin, Interpolation, Relative to CF/RL).

Remote Command:

Example:

Notes:

Initial S/W Revision:

:CALCulate:LLINe[1]|2|3|4|5|6:COPY

LLINE1|LLINE2|LLINE3|LLINE4|LLINE5|LLINE6

:CALC:LLINE2:COPY LLINE1 !copies the data from line 1 into line 2.

Auto return to the Edit menu.

A.02.00

Build from Trace

Builds a limit using an existing trace. This command will overwrite all data in the limit. Since a straight copy would typically have hundreds or thousands of segments, the data will be approximated to better represent a limit line; small excursions whose width is less than 10 trace buckets will sometimes not be captured. Secondary parameters which are not associated with traces (Description, Associated Trace, Type, Margin, Interpolation,

Relative to CF/RL) will be unchanged.

When taking a trace in order to build a limit, it will often work well to take the trace with a resolution bandwidth wider than the expected measurement, a video bandwidth lower than the expected measurement, and with the detector set to Max Hold or Min Hold.

Note that an upper limit will be built above the trace, while a lower limit will be built below the trace. If the trace is constant, the limit should pass after being built.

Remote Command:

Example:

Notes:

Initial S/W Revision:

:CALCulate:LLINe[1]|2|3|4|5|6:BUILd

TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6

:CALC:LLIN2:BUIL TRACE1 !builds limit line 2 based on the data in trace

1. This will overwrite the data in the table editor.

Auto return to Edit menu.

A.02.00

Offset

Enters a menu which allows you to offset the limit trace by a specified frequency, time, or amplitude. The offsets will be immediately applied to the limit trace for display and failure calculation; the offset can also be applied to the points in the limit line.

Key Path

Initial S/W Revision

Meas Setup, Limits, Edit

A.02.00

183

Common Measurement Functions 1

X Offset

Offsets the limit trace by some specified frequency (for Frequency-based limit lines) or a time (for time-based limit lines).

Key Path

Remote Command

Example

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

Meas Setup, Limits, Edit, Offset

:CALCulate:LLINe[1]|2|3|4|5|6:OFFSet:X <value>

:CALCulate:LLINe[1]|2|3|4|5|6:OFFSet:X?

<value> = <freq> if Limit X-Axis Unit is Frequency, <value> = <time> if

Limit X-Axis Unit is Time

:CALC:LLIN:OFFS:X –50MHZ !sets the X axis offset to –50 MHz.

:CALC:LLIN:OFFS:UPD !will apply the X axis offset to all points in the limit line, then reset the X axis offset to zero.

0 Hz if Limit X-Axis Unit is Frequency

0 S if Limit X-Axis Unit is Time

Saved in State, survives Preset

–500 GHz

500 GHz

Determined by X axis scale.

A.02.00

Y Offset

Offsets all segments in the limit line by some specified amplitude.

Key Path

Remote Command

Example

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

Meas Setup, Limits, Edit, Offset

:CALCulate:LLINe[1]|2|3|4|5|6:OFFSet:Y <rel ampl>

:CALCulate:LLINe[1]|2|3|4|5|6:OFFSet:Y?

:CALC:LLIN:OFFS:Y –3 dB !sets the Y axis offset to –3 dB.

:CALC:LLIN:OFFSet:UPD !will apply the Y axis offset to all points in the limit line, then reset the Y axis offset to zero.

0 dB

Saved in instrument state.

-Infinity

+Infinity dB

A.02.00

184

Common Measurement Functions 1

Apply Offsets to Limit Table

Adds the X and Y offsets to each point in the limit table, then resets the X and Y offset values to zero. This has no effect on the position of the limit trace.

For example, if the X offset is –10 MHz and the Y offset is 1 dB, the values in the limit table will be updated as follows: 10 MHz will be subtracted from each X value, 1 dB will be added to each Y value. The offset values will then be reset to zero. The limit trace will not be moved and the limit table will be updated to accurately reflect the currently-displayed limit trace.

Key Path

Remote Command

Example

Meas Setup, Limits, Edit, Offset

:CALCulate:LLINe[1]|2|3|4|5|6:OFFSet:UPDate

:CALC:LLIN:OFFS:UPD !sets updates the limit table to reflect the X and Y offsets, then resets the offsets to zero.

No state

A.02.00

State Saved

Initial S/W Revision

Scale X Axis

Matches the X Axis to the selected Limit, as well as possible.

For frequency limits and a frequency-domain X-axis, sets the Start and Stop Frequency to contain the minimum and maximum Frequency of the selected Limit. The range between Start Frequency and Stop Frequency is 12.5% above the range between the minimum and maximum Frequency so that span exceeds this range by one graticule division on either side.

For time limits and a time-domain X-axis, sets the sweep time to match the maximum Time of the selected Limit.

If the domain of the selected limit does not match the domain of the X Axis, no action is taken. Standard clipping rules apply, if the value in the table is outside the allowable range for the X axis.

Key Path

Initial S/W Revision

Meas Setup, Limits, Edit

A.02.00

Delete Limit

Deletes the currently selected limit line. Pressing Delete Limit purges the data from the limit line tables.

Limit data – including secondary parameters such as description, margin value, etc. - will be cleared and returned to factory preset settings.

When this key is pressed a prompt is placed on the screen that says “Please press Enter or OK key to delete limit. Press ESC or Cancel to close this dialog.” The deletion is only performed if you press OK or Enter; if so, after the deletion, the informational message “Limit deleted” appears in the MSG line.

Key Path

Remote Command

Example

Initial S/W Revision

Meas Setup, Limits

:CALCulate:LLINe[1]|2|3|4|5|6:DELete

:CALC:LLIN2:DEL !deletes all data for limit line 2.

A.02.00

185

Common Measurement Functions 1

Test Limits

Selects whether displayed traces are tested against displayed limits (i.e. those for which Limit On/Off is set to On).

For each displayed trace for which a Limit is turned on, a message will be displayed in the upper-left corner of the graticule to notify whether the trace passes or fails the limits.

If the trace is at or within the bounds of all applicable limits and margins, the text “Trace x Pass” will be displayed in green, where x is the trace number. A separate line is used for each reported trace.

If the trace is at or within the bounds of all applicable limits, but outside the bounds of some applicable margin, the text “Trace x Fail Margin” will be displayed in amber, where x is the trace number. A separate line is used for each reported trace.

If the trace is outside the bounds of some applicable limits, the the text “Trace x Fail” will be displayed in red, where x is the trace number. A separate line is used for each reported trace.

If the trace has no enabled limits, or the trace itself is not displayed, no message is displayed for that trace.

The PASS/FAIL box in the corner of the Meas Bar is only displayed if there is at least one “Trace

Pass/Fail” indication displayed in the graticule.

If two amplitude values are entered for the same frequency, a single vertical line is the result. In this case, if an upper line is chosen, the lesser amplitude is tested. If a lower line is chosen, the greater amplitude is tested.

This command only affects the display, and has no impact on remote behavior. Limit queries over SCPI test the trace against the limit regardless of whether the trace or the limit is turned on (exception: the query :CALCulate:TRACe[1]|2|3|4|5|6:FAIL? tests only the limits that are turned on for that trace).

Key Path

Remote Command

Example

Preset

State Saved

Initial S/W Revision

Meas Setup, Limits

:CALCulate:LLINe:TEST OFF|ON|0|1

:CALCulate:LLINe:TEST?

:CALC:LLIN:TEST ON !turns on testing, and displays the results in the upper left corner.

On, not affected by Mode Preset, preset by Restore Mode Defaults.

Saved in instrument state.

A.02.00

X-Axis Unit

Selects how the limit-line segments are defined. Pressing X Axis Unit selects whether the limit lines will be entered using frequency (Freq) or sweep time (Time) to define the segments. They can be specified as a table of limit-line segments of amplitude versus frequency, or of amplitude versus time.. When the

X-Axis Unit is set to Time, a time value of zero corresponds to the start of the sweep, which is at the left edge of the graticule, and the column and softkey in the Limit Table Editor will read Time instead of

Frequency

Switching the limit-line definition between Freq and Time will erase all of the current limit lines. When

186

Common Measurement Functions 1

you do this from the front panel, a warning dialog will pop up letting you know that you are about to erase all the limit lines, and prompting you to hit “OK” if you are sure:

Changing the X Axis Unit will erase all your limit lines. Are you sure you want to do this? Press

Enter

or

OK

to proceed, or

Cancel(Esc)

to cancel.

Key Path

Remote Command

Example

Couplings

Preset

State Saved

Initial S/W Revision

Meas Setup, Limits

:CALCulate:LLINe:CONTrol:DOMain FREQuency|TIME

:CALCulate:LLINe:CONTrol:DOMain?

:CALC:LLIN:CONT:DOM FREQ !deletes all currently existing limit lines, then sets all limit lines to be specified in terms of frequency.

This affects all limit lines simultaneously, and resets all limit line data except the .wav file and email address stored in the Actions.

Freq, not affected by Mode Preset, preset by Restore Mode Defaults.

Saved in instrument state.

A.02.00

Delete All Limits

Deletes all limit lines. Pressing Delete All Limits purges the data from all limit line tables.

All limit data will be cleared and returned to factory preset settings.

When this key is pressed a prompt is placed on the screen that says “Please press Enter or OK key to delete all limits. Press ESC or Cancel to close this dialog.” The deletion is only performed if you press

OK or Enter; if so, after the deletion, the informational message “All Limits deleted” appears in the MSG line.

Key Path

Remote Command

Example

Initial S/W Revision

Meas Setup, Limits

:CALCulate:LLINe:ALL:DELete

:CALC:LLIN:ALL:DEL !deletes all data for all limit lines.

A.02.00

Limit Line Data (Remote Command Only, Backwards Compatibility)

Defines the limit line values, and destroys all existing data. Up to 200 points may be defined for each limit using the following parameters.

<x>Frequency or time values as specified by :Calculate:LLINe:CONTrol:DOMain. Units default to Hz

(for frequency) and seconds (for time).

Range: –30 Gs to +30 Gs for time limits, –3 kHz to +350 GHz for frequency limits.

<ampl>Amplitude values units default to dBm. Up to two amplitude values can be provided for each x-axis value, by repeating <x-axis> in the data list.

187

Common Measurement Functions 1

Range: –1000 dBm to +1000 dBm

<connect> connect values are either "0" or "1." A "1" means this point will be connected to the previously defined point to define the limit line. A "0" means that it is a point of discontinuity and is not connected to the preceding point. The connect value is ignored for the first point.

Remote Command:

Example:

Preset:

State Saved:

Initial S/W Revision:

:CALCulate:LLINe[1]|2|3|4|5|6:DATA <x>,<ampl>,<connect>

:CALCulate:LLINe[1]|2|3|4|5|6:DATA?

:CALC:LLIN3:DATA

1E9,–20,0,2E9,–20,1,2E9,–10,1,3E9,–10,1 describes a stair-stepped limit line.

Limit line data is cleared by Restore Mode Defaults. However, it survives shutdown/restart of the analyzer application (including power cycle)

Saved in instrument state.

A.02.00

Limit Line Fail? (Remote Command Only)

Tests a limit line against its associated trace. Returns a 0 if the trace is within the limit and margin, a 1 if the trace exceeds either the limit or the margin.

Note that this command only tests one limit line – other limit lines are not tested when executing this command. To see whether a trace passed all limits, use :CALCulate:TRACe:FAIL?.

Note this command performs the test regardless of whether the trace or the limit is turned on on the display.

Remote Command:

Example:

Initial S/W Revision:

:CALCulate:LLINe[1]|2|3|4|5|6:FAIL?

:CALC:LLIN:FAIL? returns a zero if limit line 1’s associated trace has no failure, 1 if there is a margin or limit failure.

A.02.00

Limit State (Remote Command Only, SCPI standard compatibility)

Sets or queries whether the limit line is tested. This command is identical to

:CALC:LLIN[1]|2|3|4|5|6:DISP.

Remote Command:

Example:

Couplings:

Preset:

:CALCulate:LIMit[1]|2|3|4|5|6:STATe ON|OFF|0|1

:CALCulate:LIMit[1]|2|3|4|5|6:STATe?

:CALC:LIM:STAT ON turns on limit line 1

This command is identical to :CALC:LLIN:DISP

Testing is done on all displayed limits if “Test All Limits” is ON.

Off (all limits)

188

Common Measurement Functions 1

State Saved:

Initial S/W Revision:

Saved in State.

A.02.00

Limit Line Control (Remote Command Only, SCPI standard compatibility)

Defines a list of limit line control (frequency or time) values for a given limit line. Up to 2000 points may be defined for each limit using the following parameters.

<x>Frequency or time values as specified by :Calculate:LLINe:CONTrol:DOMain. Units default to Hz

(for frequency) and seconds (for time).

Range: –30 Gs to +30 Gs for time limits, –3 kHz to +1200 GHz for frequency limits.

Note that X values may be repeated if a vertical step in the limit line is desired.

The points query returns the number of points in the control. It should match the number of points in the amplitude, that is, the number of values for the CONTrol axis and for the corresponding UPPer and/or

LOWer limit lines must be identical. If one array is larger than the other, the limit trace is built using only as much data as is contained in the smaller array.

An empty array returns not a number (9.91e+37 to a data query), 0 to a POINts query.

Remote Command:

Example:

Preset:

State Saved:

Initial S/W Revision:

:CALCulate:LIMit[1]|2|3|4|5|6:CONTrol[:DATA] <x>, <x>, …

:CALCulate:LIMit[1]|2|3|4|5|6:CONTrol[:DATA]?

:CALC:LIM:CONT 1GHz,2GHz,2GHz,3GHz

!describes the X values of a stair-stepped limit line.

Limit line data is cleared by Restore Mode Defaults.

Saved in instrument state.

A.02.00

Remote Command:

Example:

Preset:

State Saved:

Initial S/W Revision:

:CALCulate:LIMit[1]|2|3|4|5|6:CONTrol:POINts?

:CALC:LIM:CONT:POIN?

!returns the number of points in the limit line.

Limit line data is cleared by Restore Mode Defaults.

Saved in instrument state.

A.02.00

Limit Line Upper / Lower (Remote Command Only, SCPI standard compatibility)

Defines a list of amplitude values for a given limit line. Changing the number of elements in the list spectrum will automatically turn the limit line off. Using the “UPP” syntax defines an upper limit line, using the “LOW” syntax defines a lower limit line. Note that a line may not be simultaneously both upper and lower; the type of the limit line will automatically be changed as appropriate. Up to 200 points may be defined for each limit using the following parameters.

<ampl>Amplitude values units default to dBm.

189

Common Measurement Functions 1

Range: –200 dBm to +100 dBm

The points query returns the number of points in the amplitude list. It will not be possible to turn on the limit line unless the number of points in the control matches the number of points in the amplitude.

The points query returns the number of points in the amplitude list. It should match the number of points in the control, that is, the number of values for the CONTrol axis and for the corresponding UPPer and/or

LOWer limit lines must be identical. If one array is larger than the other, the limit trace is built using only as much data as is contained in the smaller array.

An empty array returns the system error “list is empty” to a data query, 0 to a POINts query.

Remote Command:

Example:

Preset:

State Saved:

Initial S/W Revision:

:CALCulate:LIMit[1]|2|3|4|5|6:UPPer[:DATA] <ampl>,

<ampl>, …

:CALCulate:LIMit[1]|2|3|4|5|6:UPPer[:DATA]?

:CALC:LIM:UPP –10, –10, –20, –20

!describes the amplitude values of an upper limit line

Limit line data is cleared by Restore Mode Defaults.

Saved in instrument state.

A.02.00

Remote Command:

Example:

Preset:

State Saved:

Initial S/W Revision:

Remote Command:

:CALCulate:LIMit[1]|2|3|4|5|6:UPPer:POINts?

:CALC:LIM:UPP:POIN? !returns the number of points in the upper limit line.

Upper Limit line data/points is cleared by Restore Mode Defaults.

Saved in instrument state.

A.02.00

Example:

Preset:

State Saved:

Initial S/W Revision:

Remote Command:

Example:

Preset:

State Saved:

:CALCulate:LIMit[1]|2|3|4|5|6:LOWer[:DATA] <ampl>, …

:CALCulate:LIMit[1]|2|3|4|5|6:LOWer[:DATA]?

:CALC:LIM:LOW –10, –10, –20, –20

!describes the amplitude values of an lower limit line

Limit line data is cleared by Restore Mode Defaults.

Saved in instrument state.

A.02.00

:CALCulate:LIMit[1]|2|3|4|5|6:LOWer:POINts?

:CALC:LIM:UPP:POIN? !returns the number of points in the lower limit line.

Limit line data/points is cleared by Restore Mode Defaults.

Saved in instrument state.

190

Common Measurement Functions 1

Initial S/W Revision: A.02.00

Limit Fail? (Remote Command Only, SCPI standard Compatibility)

Tests a limit line against its associated trace. Returns a 0 if the trace is within the limit and margin, a 1 if the trace exceeds either the limit or the margin. This command is identical to “:CALC:LLIN:FAIL?”

Note that this command only tests one limit line – other limit lines are not tested when executing this command. To see whether a trace passed all limits, use :CALCulate:TRACe:FAIL?.

Note this command performs the test regardless of whether the trace or the limit is turned on on the display.

Remote Command:

Example:

Couplings:

Initial S/W Revision:

:CALCulate:LIMit[1]|2|3|4|5|6:FAIL?

:CALC:LIM:FAIL? !returns a zero if limit line 1’s associated trace has no failure, 1 if there is a margin or limit failure.

This command is identical to :CALC:LLIN:FAIL?

A.02.00

Limit Clear (Remote Command Only, SCPI standard Compatibility)

Clears a limit line, and all associated data. This command is identical to “:CALC:LLIN:DEL”

Remote Command:

Example:

Couplings:

Initial S/W Revision:

:CALCulate:LIMit[1]|2|3|4|5|6:CLEar

:CALC:LIM2:CLE !deletes all data for limit line 2.

This command is identical to :CALC:LLIN:DEL

A.02.00

Trace Fail? (Remote Command Only)

Tests a trace against all associated limit lines. Returns a 0 if the trace is within all limits and margins, a 1 if the trace exceed either the limit or the margin. If no limits apply to the selected trace, this will automatically return a 0.

Only applies to limits that are turned on, if a Limit is off it will not be tested. If a Trace is not displaying it will still be tested, and if

Test Limits (All Limits)

is off the Trace will still be tested.

This command ignores limit lines that are assigned to other traces.

Remote Command:

Example:

Initial S/W Revision:

:CALCulate:TRACe[1]|2|3|4|5|6:FAIL?

:CALC:TRAC3:FAIL? !returns a zero if there is no failure, 1 if the trace exceeds either the limit or the margin.

A.02.00

Fixed / Relative Limit (Remote Command Only, Backwards Compatibility)

This command sets both Relative to CF and Relative to RL simultaneously for all limits. If queried, it

191

Common Measurement Functions 1

returns whether Limit Line 1 is set Relative to CF, and ignores all other fixed/relative data.

Remote Command:

Example:

Notes:

Preset:

Initial S/W Revision:

:CALCulate:LLINe:CMODe FIXed|RELative

:CALCulate:LLINe:CMODe?

:CALC:LLIN:CMOD REL !makes all limit lines relative to the center frequency and reference level.

This SCPI command is only supported for Backwards Compatibility.

PSA offers only the following softkey, which is generic to all limit lines:

Limits Fixed / Rel.

On the X-Series, this functionality is provided by a softkey which is specific to each limit line, and which provides a sub-menu with 2 softkeys (Relative to

CF / Relative to RL).

In order to be consistent with the implementation of the following new commands:

:CALCulate:LLINe[1]|2|3|4|5|6:FREQuency:CMODe:RELative ON|OFF|1|0

:CALCulate:LLINe[1]|2|3|4|5|6:FREQuency:CMODe:RELative?

and

:CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:CMODe:RELative ON|OFF|1|0

:CALCulate:LLINe[1]|2|3|4|5|6:AMPLitude:CMODe:RELative?

The :CALCulate:LLINe:CMODe? Query will returns 1 if Limit Line 1 is set

Relative to CF, and returns 0 otherwise.

Fixed

A.02.00

Merge Limit Line Data

Adds the points with the specified values to the current limit line, allowing you to merge limit line data.

Up to two amplitude values are allowed for each X value. If more than 200 points are entered to be merged, the first 200 points are merged, then an error ‘too many DATA entries’ is reported.

Remote Command:

Example:

Notes:

Preset:

Initial S/W Revision:

:CALCulate:LLINe[1]|2|3|4|5|6:DATA:MERGe

<x-axis>,<ampl>,<connected>

:CALC:LLIN1:DATA:MERG 1000000000,–20,0,2000000000,–30,1 !merges the 10GHz segment and the 20GHz segment into limit line 1. Note that the

20GHz segment will be connected to the next lower point, which may or may not be the 10GHz point.

This SCPI command is only supported for Backwards Compatibility.

Although PSA had a limit of 200 points, it is acceptable to increase that limit.

Fixed

A.02.00

192

Common Measurement Functions 1

N dB Points

Turns N dB points on and off and allows you to set the N dB value. N dB uses the selected marker. If the selected marker is not on when N dB is turned on, the selected marker turns on, as a Normal marker, at center screen, and is used by N dB.

See “N dB Points Results Query” on page 193

.

See “More Information” on page 193

.

Key Path

Remote Command

Notes

Preset

Preset

State Saved

Min

Max

Initial S/W Revision

N dB Points Results Query

Meas Setup

:CALCulate:BWIDth|BANDwidth:NDB <rel_ampl>

:CALCulate:BWIDth|BANDwidth:NDB?

:CALCulate:BWIDth|BANDwidth[:STATe] OFF|ON|0|1

:CALCulate:BWIDth|BANDwidth[:STATe]?

If the selected marker is turned

Off

it turns off N dB Points.

N DB Points is unaffected by Auto Couple

Off, 3.01 dB

OFF

Off, 3.01 dB

OFF

The on/off status and the offset value are both saved in instrument state.

–140 dB

–0.01 dB

Prior to A.02.00

Remote Command:

Example:

Notes:

Initial S/W Revision:

:CALCulate:BWIDth|BANDwidth:RESult?

:CALC:MARK:AOFF set selected marker to 1

:CALC:MARK:MAX put marker 1 on peak

:CALC:BWID ON turn on N dB for the selected marker (1)

:CALC:BWID:NDB–3.01 set the offset to –3.01 dB

:CALC:BWID:RES? Query the result

–100 returned if invalid reading

Prior to A.02.00

More Information

A marker should be placed on the peak of interest before turning on N dB points. The N dB points

193

Common Measurement Functions 1

function looks for the two points on the marker’s trace closest to the marker’s X Axis value that are N dB below the marker’s amplitude, one above and the other below the marker’s X Axis value. (That is, one point is to the right and one is to the left of the selected marker.) The selected N dB value is called the offset. The function reports the frequency difference (for frequency domain traces) or time difference

(for time domain traces) between those two points.

Each point is identified by a horizontal arrow pointing towards the marker, next to the trace. The arrows used by the N dB Points function will be as shown in the figure below (where each square represents one pixel). They point in, horizontally, at the trace below a peak, on either side of its skirts. There is one pixel between the arrow and the trace

.

N dB Points can be used to measure the bandwidth of a signal; it is commonly used in conjunction with a tracking generator to measure filter bandwidths.

In one of the common use cases, the marker is placed on a peak, and the arrows are displayed N dB down the skirt from the marker on either side of the peak. The N dB value and the frequency difference between the two arrows is displayed around the arrow as shown in the figure above. Normally this displays on the right hand arrow, but if this would place any part of the text offscreen to the right then it displays on the left arrow.

If the analyzer is unable to find data that is N dB below the marker on either side of the marker, the arrows are displayed at the indicator point of the marker, no value (---) will be displayed as the result and

–100 Hz returned remotely (see figure below):

194

Common Measurement Functions 1

Some sample N dB scenarios are shown below, to illustrate how the function works in various cases. In each case, the two-headed blue arrow represents N dB of amplitude.

1 1

1 1

195

Common Measurement Functions 1

2

1

PhNoise Opt

Selects the LO (local oscillator) phase noise behavior for various desired operating conditions.

Remote Command:

Example:

Notes:

Preset:

Dependencies:

Initial S/W Revision:

Modified at S/W Revision:

[:SENSe]:FREQuency:SYNThesis[:STATe] 1|2|3

[:SENSe]:FREQuency:SYNThesis[:STATe]?

FREQ:SYNT 2 selects optimization for best wide offset phase noise

Parameter:

1 - optimizes phase noise for small frequency offsets from the carrier.

2 - optimizes phase noise for wide frequency offsets from the carrier.

3 - optimizes LO for tuning speed

(In PXA, the local oscillator hardware provides for extra-low phase noise at the expense of some speed. In these models, the “fast tuning” option lets you go faster at the expense of some noise. In all other models, the fastest possible tuning is the same as the close-in phase noise setting; in those models, the settings for option 1 are used if option 3 is selected.)

Because this function is in Auto after preset, and because Span after preset >

314.16 kHz (see Auto rules, next section) the state of this function after Preset will be 2

Does not appear in all models. The key is blank in those models, but the

SCPI command is accepted for compatibility (although no action is taken).

Prior to A.02.00

A.02.00

196

Common Measurement Functions 1

Auto

Selects the LO (local oscillator) phase noise behavior to optimize dynamic range and speed for various instrument operating conditions.

The X-Series has two grades of LO; a high performance LO that gives the best phase noise performance; and a medium-performance LO that gives excellent performance.

In models with the high performance LO, Auto will choose:

Fast Tuning whenever Span > 44.44 MHz or RBW > 1.9 MHz

otherwise, if center frequency is < 195 kHz OR ALL of the following are true:

CF 1 MHz AND Span 1.3 MHz AND RBW 75 kHz

then Best Close in Phase Noise; otherwise, Best Wide-offset Phase Noise

In models with the medium-performance LO, Auto will choose:

Fast Tuning

whenever Span > 12.34 MHz or RBW > 250 kHz

otherwise, if center frequency is < 25 kHz OR ALL of the following are true:

CF >= 1 MHz AND Span <= 141.4 kHz AND RBW <= 5 kHz

then

Best Close in Phase Noise

; otherwise,

Best Wide-offset Phase Noise

In units whose hardware does not provide for an extra-fast tuning option, the settings for Fast Tuning are the same as Best Close-in, so in those models you will see no difference between these settings.

These rules apply whether in swept spans, zero span, or FFT spans.

Key Path

Remote Command

Example

Preset

State Saved

Initial S/W Revision

Meas Setup, PhNoise Opt

[:SENSe]:FREQuency:SYNThesis:AUTO[:STATe] OFF|ON|0|1

[:SENSe]:FREQuency:SYNThesis:AUTO[:STATe]?

FREQ:SYNT:AUTO ON

ON

Saved in instrument state.

Prior to A.02.00

Best Close-in P Noise

The LO phase noise is optimized for smaller offsets from the carrier, at the expense of phase noise farther out.

Key Path

Example

Meas Setup, PhNoise Opt

FREQ:SYNT 1

197

Common Measurement Functions 1

Couplings

Readback

The frequency below which the phase noise is optimized is model dependent:

CXA: n/a

EXA: [offset <20 kHz]

MXA: [offset <20 kHz]

PXA: [offset <140 kHz]

Close-in.

If manually selected the “Man” will be underlined. The actual frequency offset within which noise is optimized is shown with in square brackets, as this can vary depending on the hardware set in use. For example, in some analyzers this annotation appears as [offset <20 kHz]

Prior to A.02.00

Initial S/W Revision

Best Wide-offset Noise

The LO phase noise is optimized for wider offsets from the carrier. Optimization is especially improved for offsets from 70 kHz to 300 kHz. Closer offsets are compromised and the throughput of measurements

(especially remote measurements where the center frequency is changing rapidly), is reduced.

Key Path

Example

Couplings

Readback

Initial S/W Revision

Meas Setup, PhNoise Opt

FREQ:SYNT 2

The frequency below which the phase noise is optimized is model dependent:

CXA: n/a

EXA: [offset >30 kHz]

MXA: [offset >30 kHz]

PXA: [offset >160 kHz]

Wide-offset.

If manually selected the “Man” will be underlined. The actual frequency offset beyond which noise is optimized is shown with in square brackets, as this can vary depending on the hardware set in use. For example, in some analyzers this annotation appears as [offset >30 kHz]

Prior to A.02.00

Fast Tuning

In this mode, the LO behavior compromises phase noise at many offsets from the carrier in order to allow rapid measurement throughput when changing the center frequency or span. The term “fast tuning” refers to the time it takes to move the local oscillator to the start frequency and begin a sweep; this setting does not impact the actual sweep time in any way.

In this mode in PXA, the LO behavior compromises phase noise at offsets below 4 MHz in order to improve measurement throughput. The throughput is especially affected when moving the LO more than 2.5 MHz and up to 10 MHz from the stop frequency to the next start frequency.

198

Common Measurement Functions 1

(In models whose hardware does not provide for a fast tuning option, the settings for Best Close-in P

Noise are used if Fast Tuning is selected. This gives the fastest possible tuning for that hardware set.)

Key Path

Example

State Saved

Readback

Initial S/W Revision

Meas Setup, PhNoise Opt

FREQ:SYNT 3

Saved in instrument state.

Fast Tuning. Also, the “Man” must be underlined.

Prior to A.02.00

ADC Dither

Accesses the menu to control the ADC Dither function. The dither function enhances linearity for low level signals at the expense of reduced clipping-to-noise ratio. The reduced clipping-to-noise ratio results in higher noise, because we work to ensure that the clipping level of the ADC relative to the front terminals remains unchanged with the introduction of dither, and this results in reduced ADC dynamic range. So making measurements with ADC dither gives you better amplitude linearity, but turning ADC dither off gives you a lower noise floor (better sensitivity).

With dither on, the third-order distortions are usually invisible for mixer levels below –35 dBm. With dither off, these distortions can be visible, with typical power levels of –110 dBm referred to the mixer.

Detection nonlinearity can reach 1 dB for dither off at mixer levels around –70 dBm and lower, while the specified nonlinearity is many times smaller with dither on.

When ADC Dither is on, the linearity of low-level signals is improved. The enhanced linearity is mostly improved scale fidelity. The linearity improvements of dither are most significant for RBWs of 3.9 kHz and less in swept mode, and FFT widths of 4 kHz and less in FFT mode.

The increased noise due to turning dither on is most significant in low band (0 to 3.6 GHz) with IF Gain set to Low, where it can be about 0.2 dB.

Key Path

Example

Remote Command

Dependencies

Preset

Meas Setup

ADC:DITH:HIGH !Sets the ADC dither setting to High

ADC:DITH ON ! Sets the ADC dither setting to Medium

In older instruments the “Medium” key was labeled “On” and the SCPI for this setting is NOT changing.

[:SENSe]:ADC:DITHer[:STATe] OFF|ON|HIGH

[:SENSe]:ADC:DITHer[:STATe]?

In some models, the “High” parameter is not available. In some instruments, the HIGH parameter is honored and the HIGH state set, and returned to a query, but the Medium dither level is actually used.

AUTO

199

Common Measurement Functions 1

Backwards Compatibility SCPI

Initial S/W Revision

Modified at S/W Revision

The old command [:SENSe]:ADC:DITHer AUTO is aliased to

[:SENSe]:ADC:DITHer:AUTO[:STATe] ON; because of this, the

[:SENSe]:ADC:DITHer function cannot be a true Boolean, so the query,

[:SENSe]:ADC:DITHer? returns OFF or ON (not 1 or 0 like a true Boolean)

Prior to A.02.00

A.02.00

Auto

Sets the ADC dither to automatic. The analyzer then chooses the dither level according to which is most likely to be the best selection, based on other settings within the digital IF.

When in Auto, the analyzer sets the dither to Medium whenever the effective IF Gain is Low by this definition of IF Gain = Low:

• When Sweep Type = Swept, IF Gain = Low whenever Swept IF Gain is set to Low Gain, whether by autocoupling or manual selection.

• When Sweep Type = FFT, IF Gain = Low whenever FFT IF Gain is set to "Low Gain," which cannot happen by autocoupling.

Whenever the IF Gain is not low by this definition, Auto sets the dither to Off.

Key Path

Remote Command

Example

Preset

State Saved

Readback

Initial S/W Revision

Meas Setup, ADC Dither

[:SENSe]:ADC:DITHer:AUTO[:STATe] OFF|ON|0|1

[:SENSe]:ADC:DITHer:AUTO[:STATe]?

ADC:DITH:AUTO ON

ON

Saved in instrument state

The “Auto” is underlined, and the readback value is whatever setting is auto-selected

Prior to A.02.00

High (Best Log Accy)

When ADC dither is set to High, the scale fidelity is especially good, most notably the relative scale fidelity. The tradeoff is that there is a modest loss of noise floor performance, up to about a decibel.

Key Path

Example

Readback

Initial S/W Revision

Meas setup, ADC Dither

ADC:DITH:HIGH

If manually selected, the readback is High, with the “Man” underlined

A.02.00

200

Common Measurement Functions 1

Medium (Log Accy)

The Medium setting of ADC Dither (known as “On” in earlier versions of the instrument software) improves the linearity of low-level signals at the expense of some noise degradation.

Key Path

Example

Readback

Initial S/W Revision

Modified at S/W Revision

Meas setup, ADC Dither

ADC:DITH:ON

If manually selected, the readback is Medium, with the “Man” underlined

Prior to A.02.00

A.02.00

Off (Best Noise)

When ADC Dither is Off, the instrument noise floor is improved, because without the need to make room for the dither, you get a lower noise floor and better sensitivity.

Key Path

Example

Readback

Initial S/W Revision

Meas setup, ADC Dither

ADC:DITH:OFF

If manually selected, the readback is Off, with the “Man” underlined.

Prior to A.02.00

Swept IF Gain

To take full advantage of the RF dynamic range of the analyzer, there is an added switched IF amplifier with approximately 10 dB of gain. When you can turn it on without overloading the analyzer, the dynamic range is always better with it on than off. The

Swept IF Gain

key can be used to set the IF Gain function to Auto, or to High Gain (the extra 10 dB), or to Low Gain. These settings affect sensitivity and

IF overloads.

This function is only active when in Swept sweeps. In FFT sweeps, the FFT IF Gain function is used instead.

Key Path

Remote Command

Example

Notes

Meas Setup

[:SENSe]:IF:GAIN:SWEPt[:STATe] OFF|ON|0|1

[:SENSe]:IF:GAIN:SWEPt[:STATe]?

IF:GAIN:SWEP ON where ON = high gain

OFF = low gain

201

Common Measurement Functions 1

Couplings

Preset

State Saved

Readback Line

Initial S/W Revision

The ‘auto’ rules for Swept IF Gain depend on attenuation, preamp state, start and stop frequency and the setting of FFT IF Gain. Set the Swept IF Gain to

High (On) when the total input attenuation is 0 dB, the preamp is off, the start frequency is 10 MHz or more, and the FFT IF Gain is autocoupled, or manually set to Autorange, or manually set to High. Also set the Swept IF

Gain to High (On) when the total input attenuation is 2 dB or less, the preamp is on, the start frequency is 10 MHz or more, and the stop frequency is 3.6

GHz or less and the FFT IF Gain is autocoupled, or manually set to

Autorange, or manually set to High. Under all other circumstances, set the

Swept IF Gain to Low (Off).

If the sweep type is Swept, the start frequency of the instrument is less than 10

MHz, and you put Swept IF Gain in Manual On, a warning condition is generated and remains in effect as long as this condition exists. The warning message is about a possible IF overload.

As with most parameters with an AUTO state, AUTO COUPLE sets it to

Auto, and setting any specific value (for example on or off) will set the AUTO state to false.

Auto after a Preset which yields Off unless the Preamp is on.

Auto and Off after Meas Preset.

Saved in instrument state.

High Gain or Low Gain

Prior to A.02.00

Auto

Activates the auto rules for Swept IF Gain

Key Path

Remote Command

Example

Preset

Initial S/W Revision

Meas setup

[:SENSe]:IF:GAIN:SWEPt:AUTO[:STATe] OFF|ON|0|1

[:SENSe]:IF:GAIN:SWEPt:AUTO[:STATe]?

IF:GAIN:SWEP:AUTO ON

ON

Prior to A.02.00

Low Gain (Best for Large Signals)

Forces Swept IF Gain to be off.

Key Path

Example

State Saved

Readback

Meas setup, ADC Ranging

IF:GAIN:SWEP OFF

Saved in instrument state.

Low Gain

202

Common Measurement Functions 1

Initial S/W Revision

High Gain (Best Noise Level)

Forces Swept IF Gain to be on.

Prior to A.02.00

Key Path

Example

Dependencies

State Saved

Readback

Initial S/W Revision

Meas setup, ADC Ranging

IF:GAIN:SWEP ON

The High setting for

Swept IF Gain

is grayed out when

FFT IF Gain

is manually set to Low (not when Low is chosen by the auto-rules).

Saved in instrument state.

High Gain

Prior to A.02.00

FFT IF Gain

Accesses the keys to set the ranging in the digital IF when doing FFT sweeps. When in Autorange mode, the IF checks its range once for every FFT chunk, to provide the best signal to noise ratio. You can specify the range for the best FFT speed, and optimize for noise or for large signals.

When the sweep type is FFT and this function is in Autorange, the IF Gain is set ON initially for each chunk of data. The data is then acquired. If the IF overloads, then the IF Gain is set OFF and the data is re-acquired. Because of this operation, the Auto setting uses more measurement time as the instrument checks/resets its range. You can get faster measurement speed by forcing the range to either the high or low gain setting. But you must know that your measurement conditions will not overload the IF (in the high gain range) and that your signals are well above the noise floor (for the low gain range), and that the signals are not changing.

Key Path

Remote Command

Couplings

Preset

State Saved

Readback Line

Initial S/W Revision

Meas Setup

[:SENSe]:IF:GAIN:FFT[:STATe] AUTOrange|LOW|HIGH

[:SENSe]:IF:GAIN:FFT[:STATe]?

As with most parameters with an AUTO state, AUTO COUPLE sets it to

Auto, which then picks AUTOrange, and setting any specific value

(AUTOrange, LOW or HIGH) will set the AUTO state to false.

AUTOrange

Saved in instrument state.

Autorange, High Gain or Low Gain

Prior to A.02.00

Auto

Allows the instrument to pick the FFT IF Gain method as appropriate. This “Auto” state is set by the

203

Common Measurement Functions 1

Auto Couple key, and it puts it in Autorange.

Key Path

Remote Command

Example

Preset

State Saved

Backwards Compatibility SCPI

Initial S/W Revision

Meas setup

[:SENSe]:IF:GAIN:FFT:AUTO[:STATe] OFF|ON|0|1

[:SENSe]:IF:GAIN:FFT:AUTO[:STATe]?

IF:GAIN:FFT:AUTO ON

ON

Saved in instrument state.

DISPlay:WINDow[1]:TRACe:Y[:SCALe]:LOG:RANGe:AUTO

Prior to A.02.00

Autorange (Slower – Follows Signals)

Turns the ADC ranging to automatic which provides the best signal to noise ratio. Autorange is usually preferred over the manual range choices.

Key Path

Example

State Saved

Readback

Initial S/W Revision

Meas setup, FFT IF Gain

IF:GAIN:FFT AUTOrange

Saved in instrument state.

Autorange

Prior to A.02.00

Low Gain (Best for Large Signals)

Forces FFT IF Gain to be off.

Key Path

Example

State Saved

Readback

Initial S/W Revision

Meas setup, FFT IF Gain

IF:GAIN:FFT LOW

Saved in instrument state.

Low Gain

Prior to A.02.00

High Gain (Best Noise Level)

Forces FFT IF Gain to be on.

Key Path

Example

Dependencies

Meas setup, FFT IF Gain

IF:GAIN:FFT HIGH

The High setting for

FFT IF Gain

is grayed out when

Swept IF Gain

is manually set to Low (not when Low is chosen by the auto-rules).

204

Common Measurement Functions 1

State Saved

Readback

Initial S/W Revision

Saved in instrument state.

High Gain

Prior to A.02.00

Analog Demod Tune & Listen

The Analog Demod Tune & Listen key opens the Analog Demod menu which contains keys to turn the demod function on and off and select modulation type. This key only appears if the N9063A Analog

Demod personality is installed and licensed, or if Option EMC is installed and licensed.

When the function is on (set to AM, FM, or PM), the demodulated signal is fed to the analyzer’s speaker.

Muting and volume control functions are done through the standard Windows speaker volume control interface.

Key Path

Remote Command

Example

Dependencies

Preset

State Saved

Initial S/W Revision

Modified at S/W Revision

Meas Setup

[:SENSe]:DEMod AM|FM|PM|OFF

[:SENSe]:DEMod?

DEM AM turns amplitude demodulation function ON

When Tune & Listen is turned on, all active traces are forced to use the same detector.

CISPR detectors (QPD, EMI Avg, RMS Avg) and Tune & LIsten are mutually exclusive. No sound output will be heard if one of these detectors is selected.

OFF

Saved in instrument state.

Prior to A.02.00

A.02.00

AM

Pressing this key, when it is not selected, selects and activates the AM demodulation function. Pressing it a second time branches to the AM Demod menu where AM demodulation settings can be adjusted.

Key Path

State Saved

Initial S/W Revision

Meas Setup, Analog Demod Tune&Listen

Saved in instrument state.

Prior to A.02.00

Channel BW (AM Demod)

Sets the RBW setting used by the hardware during the demodulation period in nonzero spans. Note that this is a separate parameter only for the demodulation function and does not affect the RBW setting in the BW menu which is used during the normal sweep. The flat top filter type must be used during the demodulation period. A 5 kHz Video Bandwidth filter is used.

205

Common Measurement Functions 1

In Zero Span, the instrument’s RBW & VBW filters are used for the demodulation; thus, the Channel

BW (and RBW filter type) will match those of the instrument. This allows gap-free listening. The

Channel BW key is grayed out and the value displayed on the key matches the current RBW of the instrument. Upon leaving zero span, the non-zero-span setting of Channel BW is restored as well as the flattop filter type.

Key Path

Remote Command

Example

Meas Setup, Analog Demod Tune&Listen, AM

[:SENSe]:DEMod:AM:BANDwidth:CHANnel <freq>

[:SENSe]:DEMod:AM:BANDwidth:CHANnel?

DEM:AM:BAND:CHAN 200 kHz

Notes

Dependencies

Couplings

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

This key/command is grayed out in zero span.

Unavailable in zero span.

In zero span only, the value is set equal to the instrument’s current RBW value and it displays that value on the softkey, but the softkey is grayed out.

30 kHz

Saved in instrument state.

390 Hz

8 MHz

Hz

Prior to A.02.00

FM

Pressing this key, when it is not selected, selects and activates the FM demodulation function. Pressing it a second time branches to the FM Demod menu where FM demodulation settings can be adjusted.

Key Path

Example

State Saved

Initial S/W Revision

Meas Setup, Analog Demod Tune&Listen

DEM FM turns frequency demodulation function ON

Saved in instrument state.

Prior to A.02.00

Channel BW (FM Demod)

Sets the RBW setting used by the hardware during the demodulation period in nonzero spans. Note that this is a separate parameter only for the demodulation function and does not affect the RBW setting in the BW menu which is used during the normal sweep. The flat top filter type must be used during the demodulation period. A 5 kHz Video Bandwidth filter is used.

In Zero Span, the instrument’s RBW & VBW filters are used for the demodulation; thus, the Channel

BW (and RBW filter type) will match those of the instrument. This allows gap-free listening. The

Channel BW key is grayed out and the value displayed on the key matches the current RBW of the

206

Common Measurement Functions 1

instrument. Upon leaving zero span, the previous setting of Channel BW and the flattop filter type are restored.

Key Path

Remote Command

Example

Meas Setup, Analog Demod Tune&Listen, FM

[:SENSe]:DEMod:FM:BANDwidth:CHANnel <freq>

[:SENSe]:DEMod:FM:BANDwidth:CHANnel?

DEM:FM:BAND:CHAN 200 MHz

Notes

Dependencies

Couplings

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

This key / command is grayed out in zero span

Unavailable in zero span.

In zero span only, the value is set equal to the instrument’s current RBW value and it displays that value on the softkey, but the softkey is grayed out.

150 kHz

Saved in instrument state.

390 Hz

8 MHz

Hz

Prior to A.02.00

De-emphasis (FM Demod only)

The De-emphasis setting controls a single-pole filter (6 dB/octave roll off), usually to counter intentional pre-emphasis in the transmitter. When De-emphasis state is OFF the hardware digital filter is bypassed, otherwise the setting is applied

The De-emphasis softkey is only available when FM is the demod selected. It is grayed out for AM and

PM.

Key Path

Remote Command

Example

Dependencies

Preset

State Saved

Readback line

Initial S/W Revision

Meas Setup, Analog Demod Tune & Listen, FM

[:SENSe]:DEMod:FM:DEEMphasis OFF|US25|US50|US75|US750

[:SENSe]:DEMod:FM:DEEMphasis?

DEM:FM:DEEM US75

DEM:FM:DEEM?

Only available in FM. Grayed out for AM and PM.

US75 (recommended for US commercial FM 75 µs pre-emphasis)

Saved in instrument state.

1-of-N selection

Prior to A.02.00

207

Common Measurement Functions 1

Off

This setting bypasses the De-emphasis filter.

Key Path

Example

Readback

Initial S/W Revision

Meas Setup, Analog Demod Tune&Listen, FM, De-emphasis

DEM:FM:DEEM OFF

Off

Prior to A.02.00

25 µs

Sets the De-emphasis time constant to 25

μµs.

Key Path

Example

Readback

Initial S/W Revision

Meas Setup, Analog Demod Tune&Listen, FM, De-emphasis

DEM:FM:DEEM US25

25

μs

Prior to A.02.00

50 µs

Sets the De-emphasis time constant to 50 µs.

Key Path

Example

Readback

Initial S/W Revision

Meas Setup, Analog Demod Tune&Listen, FM, De-emphasis

DEM:FM:DEEM US50

50

μs

Prior to A.02.00

75 µs

Sets the De-emphasis time constant to 75 µs.

Key Path

Example

Readback

Initial S/W Revision

Meas Setup, Analog Demod Tune&Listen, FM, De-emphasis

DEM:FM:DEEM US75

75

μs

Prior to A.02.00

750 µs

Sets the De-emphasis time constant to 750 µsec.

Key Path

Example

Readback

Meas Setup, Analog Demod Tune&Listen, FM, De-emphasis

DEM:FM:DEEM US750

750

μs

208

Common Measurement Functions 1

Initial S/W Revision Prior to A.02.00

PM

Pressing this key, when it is not selected, selects and activates the PM demodulation function. Pressing it a second time branches to the PM Demod menu where PM demodulation settings can be adjusted.

Key Path

Example

State Saved

Initial S/W Revision

Meas Setup, Analog Demod Tune&Listen

DEM PM turns Phase demodulation function ON

Saved in instrument state.

Prior to A.02.00

Channel BW (PM Demod)

Sets the RBW setting used by the hardware during the demodulation period in nonzero spans. Note that this is a separate parameter only for the demodulation function and does not affect the RBW setting in the BW menu which is used during the normal sweep. The flat top filter type must be used during the demodulation period. A 5 kHz Video Bandwidth filter is used.

In Zero Span, the instrument’s RBW & VBW filters are used for the demodulation; thus, the Channel

BW (and RBW filter type) will match those of the instrument. This allows gap-free listening. The

Channel BW key is grayed out and the value displayed on the key matches the current RBW of the instrument. Upon leaving zero span, the previous setting of Channel BW and the flattop filter type are restored.

Key Path

Remote Command

Example

Meas Setup, Analog Demod Tune&Listen, M

[:SENSe]:DEMod:PM:BANDwidth:CHANnel <freq>

[:SENSe]:DEMod:PM:BANDwidth:CHANnel?

DEM:PM:BAND:CHAN 200 MHz

Notes

Dependencies

Couplings

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

This key / command is grayed out in zero span

Unavailable in zero span.

In zero span only, the value is set equal to the instrument’s current RBW value and it displays that value on the softkey, but the softkey is grayed out.

100 kHz

Saved in instrument state.

390 Hz

8 MHz

Hz

Prior to A.02.00

209

Common Measurement Functions 1

Off

Pressing this key, turns the demodulation function off.

Key Path

Example

State Saved

Initial S/W Revision

Meas Setup, Analog Demod Tune&Listen

DEM OFF turns the demodulation function OFF

Saved in instrument state.

Prior to A.02.00

Demod Time

Sets the amount of time the instrument demodulates the signal after each sweep. The demodulated signal can be heard through the speaker during demodulation. In zero span, demodulation can be performed continuously, making this parameter not applicable, hence it is grayed out in zero span.

Key Path

Remote Command

Example

Notes

Dependencies

Preset

State Saved

Min

Max

Initial S/W Revision

Meas Setup, Analog Demod Tune&Listen

[:SENSe]:DEMod:TIME <time>

[:SENSe]:DEMod:TIME?

DEM:TIME 500 ms

DEM:TIME?

This key / command is grayed out in zero span

Unavailable in zero span.

500 ms

Saved in instrument state.

2 ms

100 s

Prior to A.02.00

Demod State (Remote Command Only)

Sets or queries the state of the Analog Demod Tune and Listen function. Setting the state to ON with this command will select AM demodulation by default and activate it (turn it on).

The response to the query is determined by the current setting of [:SENSE]:DEMod AM|FM|PM|OFF.

The response will be 1 if AM, FM, PM are selected, or 0 if OFF is selected..

Remote Command:

Preset:

Initial S/W Revision:

[:SENSe]:DEMod:STATe OFF|ON|0|1

[:SENSe]:DEMod:STATe?

OFF

Prior to A.02.00

210

Common Measurement Functions 1

Noise Source

This menu allows you to turn the noise source power on or off when making manual noise figure measurements. It is included in the Spectrum Analyzer Mode as an adjunct to the full controls that are available in the Noise Figure Mode. It is only available in the Swept SA measurement of the Spectrum

Analyzer Mode.

See “More Information” on page 211 .

Key Path

Remote Command

Example

Couplings

Preset

State Saved

Range

Remote Compatibility Info

Initial S/W Revision

Meas Setup

:SOURce:NOISe:TYPE NORMal | SNS

:SOURce:NOISe:TYPE?

SOUR:NOIS:TYPE NORM

If no SNS is connected, this parameter will be set to “Normal”

When Type is set to “SNS” and the SNS is disconnected, this parameter gets bumped to “Normal”

When an SNS is not connected, the SNS type will be grayed (disabled).

Normal

Saved in instrument state.

Normal | SNS

In previous Noise Figure analysis applications, this command could optionally be preceded with the :SENSe keyword. The optional :SENSe keyword is no longer supported.

Prior to A.02.00

More Information

There are 2 types of noise sources: a Smart Noise Source (SNS), and a "Normal" noise source - e.g. 346 series. This menu allows the user to control both. The SNS has its own connector on the rear of the analyzer and when it is connected the user can then select it from the “Type” 1 of N, allowing the State parameter to then control the SNS. The "Normal" source is controlled by a BNC connector that supplies

28V. If SNS is NOT connected then the “state” parameter controls the "Normal" noise source 28V BNC port. If both are connected the “Type” parameter will determine which source the “State” parameter will control. Two sources can never be controlled together. The “SNS attached” SCPI query detailed below can be used remotely to determine if an SNS is connected. SNS functionality is limited to turning on and off only. The SNS ENR data and temperature cannot be queried, unless the Noise Figure application is installed. The SNS ENR data is issued in printed form when an SNS is purchased or can be read from the analyzer’s Noise Figure application if installed, or other Agilent noise figure instruments that support the

SNS (NFA and ESA with option 219).

When first entering the Swept SA measurement the “State” will be set to OFF and the 28v BNC drive and SNS turned off to ensure the two are in sync. When the Swept SA measurement is exited, the “State” parameter will be set to OFF and the 28v BNC and SNS drive turned off.

For making manual noise figure measurements the following setup is recommended:

211

Common Measurement Functions 1

Set the SPAN to Zero

Set attenuation to 0 dB

Set the PRE-AMP ON

Set the RBW to 4MHz

Set the Detector to AVERAGE

Set the sweep time to 16ms - sets the variance correctly for good results.

Set a Band/Interval Power Marker function and set the interval over the full width of trace i.e. Left to 0s and Right to 16ms

State

This key turns the Noise Source on and off.

Key Path

Remote Command

Example

Couplings

Preset

State Saved

Remote Compatibility Info

Initial S/W Revision

Meas Setup

:SOURce:NOISe[:STATe] ON|OFF|1|0

:SOURce:NOISe[:STATe]?

SOUR:NOIS OFF

1. If an SNS is connected, and the Type is set to SNS, this parameter turns the SNS on and off.

2. When an SNS is not connected this parameter turns the BNC 28V output on and off.

3. When the SA mode is first entered this parameter is set to OFF and the 28v drive turned OFF.

4. When the SA mode is exited this parameter is set to OFF and the 28v drive turned OFF.

OFF

Saved in instrument state.

In previous Noise Figure analysis applications, this command could optionally be preceded with the :SENSe keyword. The optional :SENSe keyword is no longer supported.

Prior to A.02.00

SNS Attached (Remote Command Only)

If an Smart Noise Source (SNS) is present this command will return 1 otherwise it will return 0.

Remote Command:

Example:

Preset:

State Saved:

:SOURce:NOISe:SNS:ATTached?

SOUR:NOIS:SNS:ATT?

OFF

No

212

Common Measurement Functions 1

Remote Compatibility Info: In previous Noise Figure analysis applications, this command could optionally be preceded with the :SENSe keyword. The optional :SENSe keyword is no longer supported.

Prior to A.02.00

Initial S/W Revision:

Meas Preset

This key returns the Meas Local variables in the Swept SA measurement to their preset values. This is the same as sending the SCPI command CONF:SAN.

The only exception is Limits On/Off, which is a persistent Meas Local variable. It will be set to Off by a

Mode Preset but not by Meas Preset.

Key Path

Initial S/W Revision

Meas Setup

Prior to A.02.00

Mode

The Mode key allows you to select the available measurement applications or “Modes”. Modes are a collection of measurement capabilities packaged together to provide an instrument personality that is specific to your measurement needs. Each application software product is ordered separately by Model

Number and must be licensed to be available. Once an instrument mode is selected, only the commands that are valid for that mode can be executed.

NOTE

Key operation can be different between modes. The information displayed in Help is about the current mode.

To access Help for a different Mode you must first exit Help (by pressing the

Cancel (Esc) key). Then select the desired mode and re-access Help.

For more information on Modes, preloading Modes, and memory requirements for Modes, see

“More

Information” on page 214

Key Path

Remote Command

Example

Notes

Front-panel key

:INSTrument[:SELect]

SA|BASIC|WCDMA|CDMA2K|EDGEGSM|PNOISE|CDMA1XEV|CWLAN|WIM

AXOFDMA|CWIMAXOFDM|VSA|VSA89601|LTE|IDEN|WIMAXFIXED|LTE

TDD|TDSCDMA|NFIGURE|ADEMOD|DVB|DTMB|ISDBT|CMMB|RLC

:INSTrument[:SELect]?

:INST SA

The available parameters are dependent upon installed and licensed applications resident in the instrument. Parameters given here are an example, specific parameters are in the individual Application.

A list of the valid mode choices is returned with the INST:CAT? Query.

213

Common Measurement Functions 1

Preset

State Saved

Initial S/W Revision

Not affected by Preset. Set to SA following Restore System Defaults, if SA is the default mode.

Saved in instrument state

Prior to A.02.00

Example

Initial S/W Revision

Notes

:INST ‘SA’

Prior to A.02.00

NOTE: The query is not a quoted string. It is an enumeration as indicated in the Instrument Select table above.

The command must be sequential: i.e. continued parsing of commands cannot proceed until the instrument select is complete and the resultant SCPI trees are available.

:INSTrument[:SELect] ‘SA’|’PNOISE’|’EDGE’|’GSM’|’BASIC’ Backwards Compatibility

SCPI

More Information

The Mode name appears on the banner after the word "Agilent" followed by the Measurement Title. For example, for the Spectrum Analyzer mode with the Swept Sa measurement running:

It is possible to specify the order in which the Modes appear in the Mode menu, using the Configure

Applications utility (System, Power On, Configure Applications). It is also possible, using the same utility, to specify a subset of the available applications to load into memory at startup time, which can significantly decrease the startup time of the analyzer. During runtime, if an application that is not loaded into memory is selected (by either pressing that applications Mode key or sending that applications

:INST:SEL command over SCPI), there will be a pause while the Application is loaded. During this pause a message box that says "Loading application, please wait…" is displayed.

Each application (Mode) that runs in the X-Series signal analyzers consumes virtual memory. The various applications consume varying amounts of virtual memory, and as more applications run, the memory consumption increases. Once an application is run, some of its memory remains allocated even when it is not running, and is not released until the analyzer program (xSA.exe) is shut down.

Agilent characterizes each Mode and assigns a memory usage quantity based on a conservative estimate.

There is a limited amount of virtual memory available to applications (note that this is virtual memory and is independent of how much physical RAM is in the instrument). The instrument keeps track of how much memory is being used by all loaded applicationss - which includes those that preloaded at startup, and all of those that have been run since startup.

When you request a Mode that is not currently loaded, the instrument looks up the memory estimate for that Mode, and adds it to the residual total for all currently loaded Modes. If there is not enough virtual memory to load the Mode, a dialog box and menu will appear that gives you four options:

1.Close and restart the analyzer program without changing your configured preloads. This may free up enough memory to load the requested Mode, depending on your configured preloads

2.Clear out all preloads and close and restart the analyzer program with only the requested application preloaded, and with that application running. This choice is guaranteed to allow you to run the requested

214

Common Measurement Functions 1

application; but you will lose your previously configured preloads. In addition, there may be little or no room for other applications, depending on the size of the requested application.

3.Bring up the Configure Applications utility in order to reconfigure the preloaded apps to make room for the applications you want to run (this will then require restarting the analyzer program with your new configuration). This is the recommended choice because it gives you full flexibility to select exactly what you want.

4.Exit the dialog box without doing anything, which means you will be unable to load the application you requested.

In each case except 4, this will cause the analyzer software to close, and you will lose all unsaved traces and results.

If you attempt to load a mode via SCPI that will exceed memory capacity, the Mode does not load and an error message is returned:

-225,"Out of memory;Insufficient resources to load Mode (mode name)" where "mode name" is the SCPI parameter for the Mode in question, e.g. SA for Spectrum Analyzer

Mode

Application Mode Number Selection (Remote Command Only)

Select the measurement mode by its mode number. The actual available choices depend upon which applications are installed in your instrument. The modes appear in this table by NSEL number, which is

not the same as their order in the Mode menu (see “Detailed List of Modes” on page 219 for the mode

order).

Mode

Spectrum Analyzer

I/Q Analyzer (Basic)

WCDMA with HSDPA/HSUPA 9 cdma2000 10

1

8

:INSTrument:NSELect

<integer>

GSM/EDGE/EDGE Evo

Phase Noise

1xEV-DO

Combined WLAN

802.16 OFDMA

(WiMAX/WiBro)

13

14

15

19

75

Combined Fixed WiMAX

Vector Signal Analyzer (VXA)

89601 VSA

LTE

81

100

101

102

:INSTrument[:SELect] <parameter>

SA

BASIC

WCDMA

CDMA2K

EDGEGSM

PNOISE

CDMA1XEV

CWLAN

WIMAXOFDMA

CWIMAXOFDM

VSA

VSA89601

LTE

215

Common Measurement Functions 1

iDEN/WiDEN/MotoTalk 103

802.16 OFDM (Fixed WiMAX) 104

LTE TDD

TD-SCDMA with HSPA/8PSK

105

211

Noise Figure

Analog Demod

DVB-T/H

DTMB

219

234

235

236

ISDB-T

CMMB

239

240

Remote Language Compatibility 266

IDEN

WIMAXFIXED

LTETDD

TDSCDMA

NFIGURE

ADEMOD

DVB

DTMB

ISDBT

CMMB

RLC

Remote Command:

Example:

Notes:

Preset:

:INSTrument:NSELect <integer>

:INSTrument:NSELect?

:INST:NSEL 1

SA mode is 1

The command must be sequential: i.e. continued parsing of commands cannot proceed until the instrument select is complete and the resultant SCPI trees are available.

Not affected by Preset. Set to default mode (1 for SA mode) following Restore

System Defaults.

Saved in state

Prior to A.02.00

State Saved:

Initial S/W Revision:

Application Mode Catalog Query (Remote Command Only)

Returns a string containing a comma separated list of names of all the installed and licensed measurement modes (applications). These names can only be used with the :INSTrument[:SELect] command.

Remote Command:

Example:

Notes:

Initial S/W Revision:

:INSTrument:CATalog?

:INST:CAT?

Query returns a quoted string of the installed and licensed modes separated with a comma. Example:

"SA,PNOISE,WCDMA"

Prior to A.02.00

216

Common Measurement Functions 1

Application Identification (Remote Commands Only)

Each entry in the Mode Menu will have a Model Number and associated information: Version, and

Options.

This information is displayed in the Show System screen. The corresponding SCPI remote commands are defined here.

Current Application Model

Returns a string that is the Model Number of the currently selected application (mode).

Remote Command:

Example:

Notes:

Preset:

State Saved:

:SYSTem:APPLication[:CURRent][:NAME]?

:SYST:APPL?

Query returns a quoted string that is the Model Number of the currently selected application (Mode). Example:

"N9060A"

String length is 6 characters.

Not affected by Preset

Not saved in state, the value will be the selected application when a Save is done.

Prior to A.02.00

Initial S/W Revision:

Current Application Revision

Returns a string that is the Revision of the currently selected application (mode).

Remote Command:

Example:

Notes:

:SYSTem:APPLication[:CURRent]:REVision?

:SYST:APPL:REV?

Query returns a quoted string that is the Revision of the currently selected application (Mode). Example:

"1.0.0.0"

String length is a maximum of 23 characters. (each numeral can be an integer

+ 3 decimal points)

Preset:

State Saved:

Initial S/W Revision:

Not affected by a Preset

Not saved in state, the value will be the selected application when a Save is done.

Prior to A.02.00

217

Common Measurement Functions 1

Current Application Options

Returns a string that is the Options list of the currently selected application (Mode).

Remote Command:

Notes:

:SYSTem:APPLication[:CURRent]:OPTion?

Query returns a quoted string that is the Option list of the currently selected application (Mode). The format is the name as the *OPT? or SYSTem:OPTion command: a comma separated list of option identifiers. Example:

"1FP,2FP"

String length is a maximum of 255 characters.

Preset:

State Saved:

Example:

Initial S/W Revision:

Not affected by a Preset

Not saved in state per se, the value will be the selected application when a

Save is invoked.

:SYST:APPL:OPT?

Prior to A.02.00

Application Identification Catalog (Remote Commands Only)

A catalog of the installed and licensed applications (Modes) can be queried for their identification.

Application Catalog number of entries

Returns the number of installed and licensed applications (Modes).

Remote Command:

Example:

Preset:

State Saved:

Initial S/W Revision:

:SYSTem:APPLication:CATalog[:NAME]:COUNt?

:SYST:APPL:CAT:COUN?

Not affected by Preset

Not saved in instrument state.

Prior to A.02.00

Application Catalog Model Numbers

Returns a list of Model Numbers for the installed and licensed applications (Modes).

Remote Command:

Example:

Notes:

:SYSTem:APPLication:CATalog[:NAME]?

:SYST:APPL:CAT?

Returned value is a quoted string of a comma separated list of Model

Numbers. Example, if SAMS and Phase Noise are installed and licensed:

"N9060A,N9068A"

String length is COUNt * 7 – 1. (7 = Model Number length + 1 for comma. –1

= no comma for the 1st entry.)

218

Common Measurement Functions 1

Preset:

State Saved:

Initial S/W Revision:

Not affected by a Preset

Not saved in instrument state.

Prior to A.02.00

Application Catalog Revision

Returns the Revision of the provided Model Number.

Remote Command:

Example:

Notes:

:SYSTem:APPLication:CATalog:REVision? <model>

:SYST:APPL:CAT:REV? ‘N9060A’

Returned value is a quoted string of revision for the provided Model Number.

The revision will be a null-string ("") if the provided Model Number is not installed and licensed. Example, if SAMS is installed and licensed:

"1.0.0.0"

Preset:

State Saved:

Initial S/W Revision:

Not affected by a Preset.

Not saved in instrument state.

Prior to A.02.00

Application Catalog Options

Returns a list of Options for the provided Model Number

Remote Command:

Example:

Notes:

Preset:

State Saved:

Initial S/W Revision:

:SYSTem:APPLication:CATalog:OPTion? <model>

:SYST:APPL:CAT:OPT? ‘N9060A’

Returned value is a quoted string of a comma separated list of Options, in the same format as *OPT? or :SYSTem:OPTion?. If the provided Model Number is not installed and licensed a null-string ("") will be returned. Example, if

SAMS is installed and licensed:

"2FP"

String length is a maximum of 255 characters.

Not affected bya Preset

Not saved in instrument state.

Prior to A.02.00

Detailed List of Modes

Spectrum Analyzer

Selects the Spectrum Analyzer mode for general purpose measurements. There are several

219

Common Measurement Functions 1

measurements available in this mode. General spectrum analysis measurements, in swept and zero span, can be done using the first key in the Meas menu, labeled Swept SA. Other measurements in the Meas

Menu are designed to perform specialized measurement tasks, including power and demod measurements.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL SA

INST:NSEL 1

Prior to A.02.00

Initial S/W Revision

IQ Analyzer (Basic)

The IQ Analyzer Mode makes general purpose frequency domain and time domain measurements. These measurements often use alternate hardware signal paths when compared with a similar measurement in the Signal Analysis Mode using the Swept SA measurement. These frequency domain and time domain measurements can be used to output I/Q data results when measuring complex modulated digital signals.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL BASIC

INST:NSEL 8

Prior to A.02.00

Initial S/W Revision

W-CDMA with HSDPA/HSUPA

Selects the W-CDMA with HSDPA/HSUPA mode for general purpose measurements of signals following this standard. There are several measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL WCDMA

INST:NSEL 9

Prior to A.02.00

Initial S/W Revision

GSM/EDGE/EDGE Evo

Selects the GSM with EDGE mode for general purpose measurements of signals following this standard.

There are several measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If

220

Common Measurement Functions 1

it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Example

Initial S/W Revision

Key Path

Modified at S/W Revision

INST:SEL EDGEGSM

INST:NSEL 13

Prior to A.02.00

Mode

A.02.00

802.16 OFDMA (WiMAX/WiBro)

Selects the OFDMA mode for general purpose measurements of WiMAX signals. There are several measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL WIMAXOFDMA

INST:NSEL 75

Prior to A.02.00

Initial S/W Revision

Vector Signal Analyzer (VXA)

The 89601X Vector Signal Analyzer provides vector-signal analysis measurement capability. It provides

3 main measurements that allow you to measure the signal quality of all varieties of RF modulation:

• Digital Demodulation

• Analog Demodulation

• Vector Signal Analysis

There are advanced modulation analysis and troubleshooting capabilities including the following communications formats: AM, FM, PM, WiMAX, W-CDMA, 2G, 3G, 3.5G, WLAN, digital video, and more. It also provides standard-specific measurements for analysis of iDEN, WiDEN, and MotoTalk signals with Option H09.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Initial S/W Revision

Mode

INST:SEL VSA

INST:NSEL 100

Prior to A.02.00

221

Common Measurement Functions 1

Phase Noise

The Phase Noise mode provides pre-configured measurements for making general purpose measurements of device phase noise.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Initial S/W Revision

Mode

INST:SEL PNOISE or

INST:NSEL 14

Prior to A.02.00

Noise Figure

The Noise Figure mode provides pre-configured measurements for making general purpose measurements of device noise figure.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Initial S/W Revision

Mode

INST:SEL NFIGURE

Or

INST:NSEL 219

Prior to A.02.00

Analog Demod

Selects the Analog Demod mode for making measurements of AM, FM and phase modulated signals.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL ADEMOD

INST:NSEL 234

Prior to A.02.00

Initial S/W Revision

TD-SCDMA with HSPA/8PSK

Selects the TD-SCDMA mode for general purpose measurements of signals following this standard.

There are several measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If

222

Common Measurement Functions 1

it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL TDSCDMA

INST:NSEL 211

Prior to A.02.00

Initial S/W Revision

cdma2000

Selects the cdma2000 mode for general purpose measurements of signals following this standard. There are several measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL CDMA2K

INST:NSEL 10

Prior to A.02.00

Initial S/W Revision

1xEV-DO

Selects the 1xEV-DO mode for general purpose measurements of signals following this standard. There are several measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL CDMA1XEV

INST:NSEL 15

Prior to A.02.00

Initial S/W Revision

LTE

Selects the LTE mode for general purpose measurements of signals following the LTE FDD standard.

There are several measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL LTE

INST:NSEL 102

223

Common Measurement Functions 1

Initial S/W Revision Prior to A.02.00

LTE TDD

Selects the LTE TDD mode for general purpose measurements of signals following the LTE TDD standard. There are several measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Initial S/W Revision

Mode

INST:SEL LTETDD

INST:NSEL 105

A.03.00

DVB-T/H

Selects the DVB-T/H mode for measurements of digital video signals using this format. There are several power and demod measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Initial S/W Revision

Mode

INST:SEL DVB

INST:NSEL 235

A.02.00

DTMB

Selects the DTMB mode for measurements of digital video signals using this format. There are several power and demod measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Initial S/W Revision

Mode

INST:SEL DTMB

INST:NSEL 236

A.02.00

ISDB-T

Selects the ISDB-T mode for measurements of digital video signals using this format. There are several power and demod measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If

224

Common Measurement Functions 1

it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL ISDBT

INST:NSEL 239

A.03.00

Initial S/W Revision

CMMB

Selects the CMMB mode for measurements of digital video signals using this format. There are several power and demod measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL CMMB

INST:NSEL 240

A.03.00

Initial S/W Revision

Combined WLAN

Selects the CWLAN mode for general purpose measurements of signals following this standard. There are several measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL CWLAN

INST:NSEL 19

A.02.00

Initial S/W Revision

Combined Fixed WiMAX

Selects the Combined Fixed WiMAX mode for general purpose measurements of signals following this standard. There are several measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL CWIMAXOFDM

INST:NSEL 81

225

Common Measurement Functions 1

Initial S/W Revision A.02.00

802.16 OFDM (Fixed WiMAX)

Selects the 802.16 OFDM (Fixed WiMAX) mode. This mode allows modulation quality measurements of signals that comply with IEEE 802.16a–2003 and IEEE 802.16–2004 standards, with flexibility to measure nonstandard OFDM formats. Along with the typical digital demodulation measurement results, several additional 802.16 OFDM unique trace data formats and numeric error data results provide enhanced data analysis.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL WIMAXFIXED

INST:NSEL 104

A.02.00

Initial S/W Revision

iDEN/WiDEN/MOTOTalk

Selects the iDEN/WiDEN/MOTOTalk mode for general purpose measurements of iDEN and iDEN-related signals. There are several measurements available in this mode.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Example

Mode

INST:SEL IDEN

INST:NSEL 103

A.02.00

Initial S/W Revision

Remote Language Compatibility

The Remote Language Compatibility (RLC) mode provides remote command backwards compatibility for the 8560 series of spectrum analyzers, known as legacy spectrum analyzers.

NOTE

After changing into or out of RLC mode, allow a 1 second delay before sending any subsequent commands.

If you are using the Help feature, this mode must be currently active to access its detailed information. If it is not active, exit the Help feature (Esc key), select the mode, and re-access Help.

Key Path

Mode

226

Common Measurement Functions 1

Example

Initial S/W Revision

INST:SEL RLC

Or

INST:NSEL 266

Prior to A.02.00

89601 VSA

Selecting the 89601 VSA mode will start the 89600-Series VSA software application. The 89600 VSA software is powerful, PC-based software, offering the industry's most sophisticated general purpose and standards specific signal evaluation and troubleshooting tools for the R&D engineer. Reach deeper into signals, gather more data on signal problems, and gain greater insight.

• Over 30 general-purpose analog and digital demodulators ranging from 2FSK to 1024QAM

• Standards specific modulation analysis including:

• Cell: GSM, cdma2000, WCDMA, TD-SCDMA and more

• Wireless networking: 802.11a/b/g, 802.11n, 802.16 WiMAX (fixed/mobile), UWB

• RFID

• Digital satellite video and other satellite signals, radar, LMDS

• Up to 400K bin FFT, for the highest resolution spectrum analysis

• A full suite of time domain analysis tools, including signal capture and playback, time gating, and

CCDF measurements

• Six simultaneous trace displays and the industry's most complete set of marker functions

• Easy-to-use Microsoft ® Windows ® graphical user interface

For more information see the Agilent 89600 Series VSA web site at www.agilent.com/find/89600

To learn more about how to use the 89600 VSA running in the X-Series, after the 89600 VSA application is running, open the 89600 VSA Help and open the "About Agilent X-Series Signal

Analyzers (MXA/EXA) with 89600-Series Software" help topic.

Key Path

Example

Initial S/W Revision

Mode

INST:SEL VSA89601

INST:NSEL 101

Prior to A.02.00

Global Settings

Opens up a menu that allows you to switch certain Meas Global parameters to a Mode Global state.

These switches apply to all Modes that support global settings. No matter what Mode you are in when you set the “Global Center Frequency” switch to on, it applies to all Modes that support Global Settings.

Key Path

Mode Setup

227

Common Measurement Functions 1

Initial S/W Revision Prior to A.02.00

Global Center Freq

The software maintains a Mode Global value called “Global Center Freq”.

When the

Global Center Freq

key is switched to

On

in any mode, the current mode’s center frequency is copied into the Global Center Frequency, and from then on all modes which support global settings use the Global Center Frequency. So you can switch between any of these modes and the Center Freq will remain unchanged.

Adjusting the Center Freq of any mode which supports Global Settings, while

Global Center Freq

is

On

, will modify the Global Center Frequency.

When

Global Center Freq

is turned

Off

, the Center Freq of the current mode is unchanged, but now the

Center Freq of each mode is once again independent.

When

Mode Preset

is pressed while

Global Center Freq

is

On

, the Global Center Freq is preset to the preset Center Freq of the current mode.

This function is reset to Off when the Restore Defaults key is pressed in the Global Settings menu, or when

System

,

Restore Defaults

,

All Modes

is pressed.

Key Path

Scope

Remote Command

Example

Preset

Range

Initial S/W Revision

Mode Setup, Global Settings

Mode Global

:INSTrument:COUPle:FREQuency:CENTer ALL|NONE

:INSTrument:COUPle:FREQuency:CENTer?

INST:COUP:FREQ:CENT ALL

INST:COUP:FREQ:CENT?

Set to Off on Global Settings, Restore Defaults and System, Restore Defaults, All Modes

On|Off

Prior to A.02.00

Remote Command:

Preset:

Initial S/W Revision:

:GLOBal:FREQuency:CENTer[:STATe] 1|0|ON|OFF

:GLOBal:FREQuency:CENTer[:STATe]?

Off

Prior to A.02.00

Restore Defaults

This key resets all of the functions in the Global Settings menu to Off. This also occurs when

System

,

Restore Defaults

,

All Modes

is pressed.

Key Path

Mode Setup, Global Settings

228

Common Measurement Functions 1

Remote Command

Example

:INSTrument:COUPle:DEFault

INST:COUP:DEF

Backwards Compatibility SCPI :GLOBal:DEFault

Initial S/W Revision Prior to A.02.00

Mode Setup

The Mode Setup menu contains setup functions that are global across the entire Mode. These functions are independent of which measurement is currently running - they are global to all measurements in the mode, or "Meas Global." The Mode Setup functions are not the only Meas Global functions in the analyzer; for example, the Trigger Setup functions are Meas Global, and there are even Mode Global functions (that is, the same for all Modes) in the Input/Output menu, but the fact that they are all Meas

Global is a distinguishing characteristic of the Mode Setup functions.

The Mode Setup menu also contains the

Restore Mode Defaults

key. Most Meas Global functions are restored to their preset values by

Mode Preset

, however some variables are more persistent and are not preset until the

Restore Mode Defaults

key is pressed.

There are also a few Meas Global variables (for example, Global Center Frequency) that can be switched to be Mode Global, that is, the same for all modes. The keys under the Global Settings key control whether these variables are Mode Global or not.

In the Spectrum Analyzer mode, the Mode Setup functions include which radio standard and/or EMC standard is in use and how it is configured. A set of CISPR EMC presets is available as well.

The EMC keys require Option EMC to be installed and licensed.

Key Path

Initial S/W Revision

Front-panel key

Prior to A.02.00

Radio Standard

Allows you to specify the radio standard to be used. Spectrum Analyzer mode supports many radio standards. Youcan select the desired radio standard using the

Radio Std

key. However, not every measurement in the Spectrum Analyzer mode is available with every standard. The chart below describes which measurements are available with each radio standard.

229

Common Measurement Functions 1

Swept SA

Channel

Power

Occupied

BW

ACP

Power Stat

CCDF

Burst Power

Spurious

Emission

Spectrum

Emission Mask

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X X

X

X

X X X

X

X

X X

X X X X

X

X

X

X

X X

X

X

X

X

X

X

X

X

X

Key Path

Scope

Remote Command

Example

Couplings

State Saved

Range

Initial S/W Revision

Mode Setup

Meas Global

[:SENSe]:RADio:STANdard[:SELect]

NONE|JSTD|IS95a|IS97D|IS98D|GSM|W3GPP|CDMA2000MC1|C2000

1X|NADC|PDC|BLUEtooth|TETRa|WL802DOT11A|WL802DOT11B|WL8

02DOT11G|HIPERLAN2|DVBTLSN|DVBTGPN|DVBTIPN|FCC15|SDMBSE

|UWBINDOOR

[:SENSe]:RADio:STANdard[:SELect]?

RAD:STAN NONE

RAD:STAN?

By changing the radio standard, the measurement parameters will be automatically set to an appropriate default value.

Saved in instrument state

None|IS–95A|J-STD–008|IS–97D/98D|GSM/EDGE|3GPP

W-CDMA|cdma2000

1x|NADC|PDC|Bluetooth|TETRA|W-LAN|DVB-T|FCC Part 15 Subpart

F|S-DMB System E|UWB Indoor

Prior to A.02.00

Device

Allows you to specify the device to be used. This key appears in the Setup menu of most of the Radio

Stds. It is a global setting that affects the Device selection, between Mobile (MS) and Base Station

(BTS) settings, for all relevant Power Suite measurements.

Key Path

Scope

Mode Setup, Radio Std Setup

Meas Global

230

Common Measurement Functions 1

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

[:SENSe]:RADio:STANdard:DEVice BTS|MS

[:SENSe]:RADio:STANdard:DEVice?

RAD:STAN:DEV MS

RAD:STAN:DEV?

BTS

Saved in instrument state

BTS|MS

Prior to A.02.00

IS–95A

Sets the specific parameters for the selected measurement appropriate for industry standard IS–95A. For the available measurements with this radio standard, see the chart in the

“Radio Standard” on page 229

section.

Key Path

Initial S/W Revision

Mode Setup, Radio Std

Prior to A.02.00

Device

Operation of this key is identical for many locations in this measurement. For details about this key, see

“Device” on page 230 .

J-STD–008

Sets the specific parameters for the selected measurement appropriate for industry standard J-STD–008.

For the available measurements with this radio standard, see the chart in the

“Radio Standard” on page 229

section.

Key Path

Initial S/W Revision

Mode Setup, Radio Std

Prior to A.02.00

Device

Operation of this key is identical for many locations in this measurement. For details about this key, see

“Device” on page 230 .

IS–97D/98D

Sets the specific parameters for the selected measurement appropriate for industry standard IS–97D/98D.

For the available measurements with this radio standard, see the chart in the

“Radio Standard” on page 229

section.

Key Path

Initial S/W Revision

Mode Setup, Radio Std

Prior to A.02.00

231

Common Measurement Functions 1

Band Class

This function is only available when you have selected the standard: IS–97D/98D. It enables you to select the band class.

Key Path

Scope

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Mode Setup, Radio Std Setup

Meas Global

[:SENSe]:RADio:STANdard:BAND:CLASs BC0|BC1

[:SENSe]:RADio:STANdard:BAND:CLASs?

RAD:STAN:BAND:CLAS BC0

RAD:STAN:BAND:CLAS?

BC0

Saved in instrument state

0 (800 MHz Band)|1 (1900 MHz Band)

Prior to A.02.00

Device

Operation of this key is identical for many locations in this measurement. For details about this key, see

“Device” on page 230

.

GSM/EDGE

Sets the specific parameters for the selected measurement appropriate for industry standard GSM/EDGE.

For the available measurements with this radio standard, see the chart in the “Radio Standard” on page 229 section.

Key Path

Initial S/W Revision

Mode Setup, Radio Std

Prior to A.02.00

Device

Operation of this key is identical for many locations in this measurement. For details about this key, see

“Device” on page 230

.

3GPP W-CDMA

Sets the specific parameters for the selected measurement appropriate for industry standard 3GPP

W-CDMA. For the available measurements with this radio standard, see the chart in the “Radio

Standard” on page 229

section.

Key Path

Initial S/W Revision

Mode Setup, Radio Std

Prior to A.02.00

Device

Operation of this key is identical for many locations in this measurement. For details about this key, see

“Device”

232

Common Measurement Functions 1

on page 230 .

Cdma2000 1x

Sets the specific parameters for the selected measurement appropriate for industry standard

cdma2000–1x. For the available measurements with this radio standard, see the chart in the “Radio

Standard” on page 229 section.

Key Path

Initial S/W Revision

Mode Setup, Radio Std

Prior to A.02.00

Device

Operation of this key is identical for many locations in this measurement. For details about this key, see

“Device” on page 230 .

NADC

Sets the specific parameters for the selected measurement appropriate for industry standard NADC. For the available measurements with this radio standard, see the chart in the

“Radio Standard” on page 229

section.

Key Path

Initial S/W Revision

Mode Setup, Radio Std

Prior to A.02.00

Device

Operation of this key is identical for many locations in this measurement. For details about this key, see

“Device” on page 230 .

PDC

Sets the specific parameters for the selected measurement appropriate for industry standard PDC. For the available measurements with this radio standard, see the chart in the

“Radio Standard” on page 229

section.

Key Path

Initial S/W Revision

Mode Setup, Radio Std

Prior to A.02.00

Device

Operation of this key is identical for many locations in this measurement. For details about this key, see

“Device” on page 230 .

Bluetooth

Sets the specific parameters for the selected measurement appropriate for industry standard Bluetooth™.

For the available measurements with this radio standard, see the chart in the Radio Standard section.

Key Path

Scope

Mode Setup, Radio Std

Meas Global

233

Common Measurement Functions 1

Remote Command

Example

Notes

Preset

State Saved

Range

Initial S/W Revision

[:SENSe]:RADio:STANdard:PACKet DH1|DH3|DH5

[:SENSe]:RADio:STANdard:PACKet?

RAD:STAN:PACK DH1

RAD:STAN:PACK?

The packet length is

DH1 -> 366 µs

DH3 -> 1622 µs

DH5 -> 2870 µs

DH1

Saved in instrument state

DH1|DH3|DH5

Prior to A.02.00

TETRA

Sets the specific parameters for the selected measurement appropriate for industry standard TETRA. For

the available measurements with this radio standard, see the chart in the “Radio Standard” on page 229

section.

Key Path

Initial S/W Revision

Mode Setup, Radio Std

Prior to A.02.00

Device

Operation of this key is identical for many locations in this measurement. For details about this key, see

“Device” on page 230

.

W-LAN

Accesses the W-LAN radio standards key menu to enable you to select a W-LAN standard. Selecting a

W-LAN standard modifies spectrum analyzer settings for the measurement activated under the Meas menu. For the available measurements with this radio standard, see the chart in the

“Radio Standard” on page 229 section.

Key Path

Range

Initial S/W Revision

Mode Setup, Radio Std

802.11a|802.11b|802.11g|HiperLAN/2

Prior to A.02.00

DVB-T

Accesses the DVB-T key menu to enable you to select a DVB-T mask filtering standard. Selecting a

DVB-T filtering standard modifies spectrum analyzer settings for the measurement activated under the

Meas menu. For the available measurements with this radio standard, see the chart in the

“Radio

234

Common Measurement Functions 1

Standard” on page 229 section.

Key Path

Range

Initial S/W Revision

Mode Setup, Radio Std

L/SECAM/NICAM|G/PAL/NICAM|I/PAL/NICAM

Prior to A.02.00

FCC Part 15 Subpart F

Sets the specific parameters for the selected measurement appropriate for unlicensed devices. For the available measurements with this radio standard, see the chart in the

“Radio Standard” on page 229

section.

Key Path

Initial S/W Revision

Mode Setup, Radio Std

Prior to A.02.00

S-DMB System E

Sets the specific parameters for the selected measurement appropriate for industry standard System E.

For the available measurements with this radio standard, see the chart in the

“Radio Standard” on page 229

section.

Key Path

Initial S/W Revision

Mode Setup, Radio Std

Prior to A.02.00

UWB Indoor

Sets the specific parameters for the selected measurement appropriate for UWB Indoor standard. For the available measurements with this radio standard, see the chart in the

“Radio Standard” on page 229

section.

Key Path

Initial S/W Revision

Mode Setup, Radio Std

Prior to A.02.00

Enable Non-Std Measurements

Allows you to specify whether all measurements and radio standards are enabled or not. In default,

Enable All Measurements is set to No, so youcan select only the valid combination of preset available standard and measurement. Any measurement or standard that make the combination that haveno valid preset value are grayed out. When Enable Non-Std Measurements is set to Yes, all measurements and standard selections are enabled so that youcan choose any.

If you select an unavailable measurement or unavailable radio standardusing the Enable Non-Std

Measurement key, the measurement results may not conform to the selected standard.

Key Path

Scope

Mode Setup

Meas Global

235

Common Measurement Functions 1

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

[:SENSe]:RADio:STANdard:EAMeas YES|NO

[:SENSe]:RADio:STANdard:EAMeas?

RAD:STAN:EAM YES

RAD:STAN:EAM?

NO

Saved in instrument state

Yes|No

Prior to A.02.00

EMC Standard

Only appears with Option EMC installed and licensed.

This menu allows youto select None (no EMI standard), CISPR (CISPR 16–1–1), and MIL

(MIL–461A). Each standard has a unique way of determining the couplings between detectors and

RBWs, as well as its own set of available RBW’s.

Note that Auto Couple will have no effect on the EMC Standard setting.

Key Path

Scope

Remote Command

Example

Dependencies

Couplings

State Saved

Initial S/W Revision

Mode Setup

Meas Global

[:SENSe]:EMC:STANdard[:SELect] NONE|CISPr|MIL

[:SENSe]:EMC:STANdard[:SELect]?

:EMC:STAN CISP

When the

EMC Standard

changes to

CISPR

or

MIL

, the

RBW Control

key is grayed out. The

Filter Type

is then always Gaussian; the

Filter BW

is chosen as appropriate for the filter and the standard.

When the EMC Standard changes to None, the Filter Type is set to Gaussian and the Filter BW is set to –3 dB.

Only appears with Option EMC installed and licensed. If not, the SCPI command generates an error.

The auto rules for detector select Peak for any trace in Auto when the EMI

Standard is CISPR or MIL.

Choosing a CISPR detector or CISPR presets automatically picks the CISPR

Standard, however switching from a CISPR detector has no impact on EMC

Standard.

Saved in instrument state

A.02.00

236

Common Measurement Functions 1

Remote Command:

Notes:

Dependencies:

[:SENSe]:BANDwidth|BWIDth:[:RESolution]:MODE

EMI|SAN|OFF

[:SENSe]:BANDwidth|BWIDth:[:RESolution]:MODE?

This command is mapped to the EMC:STANdard command with the following mappings: EMI=>CISPr, SAN =>None, and if the legacy command comes in with the OFF parameter, it sets EMC Standard to None and Res BW to Manual . The query returns “OFF” if Res BW in Manual, otherwise “EMI” if EMC Standard is CISPR or MIL, and “SAN” if EMC

Standard is None

Only appears with Option EMC installed and licensed. If not, the SCPI command generates an error.

SAN

A.02.00

Preset:

Initial S/W Revision:

CISPR presets

Only appears with Option EMC installed and licensed.

This key lets you easily set up the analyzer for CISPR measurements.

This topic contains the following sections:

“Band Setup” on page 237

“Sweep Points in Band E” on page 238

“Number of points in band E” on page 239

Key Path

Remote Command

Example

Couplings

Band Setup Band A

Mode Setup, CISPR presets

[:SENSe]:FREQuency:CISPr:BAND A|B|C|CD|D|E

FREQ:CISPR:BAND A activates the CISPR preset for Band A

Selecting a CISPR presetsets the EMI Standard to CISPR, performs an autocouple all, and sets the Y Axis Unit to dB

μV (unless dBuV is grayed out, in which case it will leave the Y Axis Unit unaffected).

A.02.00

Initial S/W Revision

Band Setup

The number of sweep points for each band is roughly calculated by the formula 2*(Stop Frequency-Start

Frequency)/RBW, so that you get two points for every RBW width. This number is increased as necessary to make it an odd integer, so that you always end up with an odd number of sweep points. This is desirable so that you always have a sweep point at the Center Freq.

Band B Band C Band D Band C&D Band E

237

Common Measurement Functions 1

Start

Frequency

Stop

Frequency

9kHz

150kHz

150kHz

30MHz

30MHz

300MHz

300MHz

1GHz

30MHz

1GHz

1GHz

Max freq of analyzer or 18

GHz, whichever is lower

See below Sweep Point 1411 6635 4501 11667 16167

The table above is based on the fact that the Res BW autocouples to the center frequency when in the

CISPR EMC standard as follows:

Center Frequency

<150 kHz

150 kHz to 30 MHz

>30 MHz to 1 GHz

>1 GHz

RBW

200 Hz

9 kHz

120 kHz

1 MHz

Sweep Points in Band E

Note that the Res BW will be 1 MHz in band E. The number of sweep points for band E is dependent on the maximum frequency of the analyzer. The formula above gives the following values for Band E:

238

Common Measurement Functions 1

Number of points in band E

Option Max

Analyzer

Freq

(nominal)

3.0 Ghz 503 (3.0 GHz models)

503 (3.6 GHz models 3.6 Ghz

507 (7 GHz models) 7.0 Ghz

507 (7.5 GHz models)

7.5 Ghz

508

513

526

8.4 GHz

Width of

Band E

2.0 GHz

2.6 GHz

6.0 GHz

6.5 GHz

7.4 GHz

13.2 GHz 12.2 GHz

26.5 GHz 17 GHz

Number of

Points

4001

5201

12001

13001

14801

24401

34001

Noise Reduction

Noise Reduction accesses a menu for configuring the noise compensation of the instrument. This menu only appears in models that support Noise Reduction.

Key Path

Initial S/W Revision

Mode Setup

A.04.00

Noise Floor Extension

Turns on the

Noise Floor Extension

function. When this function is On, the expected noise power of the analyzer (derived from a factory calibration) is subtracted from the trace data. When

Noise Floor

Extension

is On, it will usually reduce the apparent noise level by about 10 dB in low band, and 8 dB in high band (>~3.6 GHz).

Noise Floor Extension

works with any RBW, VBW, detector, any setting of Average Type, any amount of trace averaging, and any signal type. It is ineffective when the trace is not smoothed (smoothing processes include narrow VBWs, trace averaging, and long sweep times with the detector set to Average or Peak). It works best with extreme amounts of smoothing. It works best with the average detector, with the Average Type set to Power.

In those cases where the cancellation is ineffective, it nonetheless has no undesirable side-effects. There is no significant speed impact to having

Noise Floor Extension

on.

The best accuracy is achieved when substantial smoothing occurs in each point before trace averaging.

Thus, when using the average detector, results are better with long sweep times and fewer trace averages.

When using the sample detector, the VBW filter should be set narrow with less trace averaging, instead of a wide VBW filter with more trace averaging.

239

Common Measurement Functions 1

See

“More Information” on page 240

Remote Command

Example

Initial S/W Revision

Key Path

Scope

Dependencies

Preset

State Saved

[:SENSe]:CORRection:NOISe:FLOor ON|OFF|1|0

[:SENSe]:CORRection:NOISe:FLOor?

CORR:NOIS:FLO ON

A.04.00

Mode Setup, Noise Reduction

Meas Global

In models that do not support Noise Floor Extension, the SCPI command will be accepted without error but will have no effect.

Unaffected by

Mode Preset

. Turned off by

Restore Mode Defaults

.

No

More Information

The analyzer is characterized in the factory (or during a field calibration) with a model of the noise, referred to the input mixer, versus frequency in each band and path combination. Bands are 0 (low band) and 1 through 4 (high band) in a 26.5 GHz instrument, for example. Paths include normal paths, preamp paths, the electronic attenuator, etc.

In most band/path combinations, the noise can be well characterized based on just two parameters and the analyzer frequency response before compensation for frequency-dependent losses.

After the noise density at the input mixer is estimated, the effects of the input attenuator, RBW, detector, etc. are computed to get the estimated input-port-referred noise level.

In the simplest case, the measured power (signal plus analyzer noise) in each display point (bucket) is compensated by subtracting the estimated noise power, leaving just the signal power. This is the operation when the detector is Average and the Average Type is set to Power.

In other cases, operation is often not quite as good but still highly effective. With peak detection, the noise floor is estimated based on the RBW and the duration of the bucket using the same equations used in the noise marker function. The voltage of the noise is subtracted from the voltage of the observed signal-plus-noise measurement to compute the estimated signal voltage. The peak detector is one example of processing that varies with detector to give good estimates of the signal level without the analyzer noise.

For best operation, the average detector and the power scale are recommended, as already stated. Peak detection for pulsed-RF can still give excellent effectiveness. FFT analysis does not work well, and does not do NFE well, with pulsed-RF signals, so this combination is not recommended. Negative peak detection is not very useful, either. Sample detection works well, but is never better than the average detector because it doesn’t smooth as well. The Normal detector is a combination of peak and negative peak behaviors, and works about as well as these.

For best operation, extreme smoothing is desirable, as already stated. Using narrow VBWs works well, but using very long bucket durations and the average detector works best. Reducing the number of trace points will make the buckets longer.

240

Common Measurement Functions 1

For best operation, the power scale (Average Type = Power) is optimum. When making CW measurements in the presence of noise without NFE, averaging on the decibel scale has the advantage of reducing the effect of noise. When using NFE, the NFE does an even better job than using the log scale ever could. Using NFE with the log scale is not synergistic, though; NFE with the power scale works a little better than NFE with log averaging type.

Restore Mode Defaults

Resets the state for the currently active mode by resetting the mode persistent settings to their factory default values, clearing mode data and by performing a Mode Preset

See “Restore Mode Defaults” on page 134 under “Preset” in the System Functions section, for more

information.

Global Settings

Opens a menu that allows you to switch certain Meas Global parameters to a Mode Global state.

See “Global Settings” on page 227 under Mode Setup in the Common Measurement Functions section

for more information.

Peak Search

Pressing the Peak Search key displays the Peak Search menu and places the selected marker on the trace point with the maximum y-axis value for that marker’s trace. The Peak Search features allow you to define specific search criteria to determine which signals can be considered peaks, excluding unwanted signals from the search.

See “More Information” on page 241

.

Remote Command:

Example:

Notes:

Initial S/W Revision:

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MAXimum

CALC:MARK2:MAX performs a peak search using marker 2.

CALC:MARK2:Y? queries the marker amplitude (Y-axis) value for marker 2.

CALC:MARK2:X? queries the marker frequency or time (X-axis) value for marker 2.

SYST:ERR? can be used to query the errors to determine if a peak is found.

The error –200 will be returned after an unsuccessful search.

Sending this command selects the subopcoded marker.

Prior to A.02.00

More Information

If

Same as “Next Peak” Criteria

is selected, and either

Pk Excursion

or

Pk Threshold

are on, a signal must meet those criteria. If no valid peak is found, a message is generated. And then the marker is not moved.When

Highest Peak

is on, or both

Pk Excursion

and

Pk Threshold

are off, the marker is always placed at the point on the trace with the maximum y-axis value, even if that point is on the very edge of the trace (exception: negative frequencies and signals close to the LO are not searched at all.

Pressing Peak Search with the selected marker off causes the selected marker to be set to

Normal

at the

241

Common Measurement Functions 1

center of the screen, then a peak search is immediately performed.

Pressing the front panel Peak Search key always does a peak search. Occasionally, you may need to get to the Peak Search menu key functions without doing a peak search. You can do this by first accessing the Peak Search menu. Then go to the other menus that you need to access. Finally, you can get back to the Peak Search key menu by using the front panel Return key and pressing it as many times as required to navigate back through the previously accessed menus until you get back to the Peak Search menu.

Next Peak

Pressing Next Peak moves the selected marker to the peak that has the next highest amplitude less than the marker’s current value. Only peaks which meet all enabled peak criteria are considered. If there is no valid peak lower than the current marker position, an error is generated and the marker is not moved.

If the selected marker was off, then it is turned on as a normal marker and a peak search is performed.

Key Path

Remote Command

Example

Notes

State Saved

Initial S/W Revision

Peak Search

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MAXimum:N

EXT

CALC:MARK2:MAX:NEXT !Selects marker 2 and moves it to the peak that is closest in amplitude to the current peak, but the next lower value.

Sending this command selects the subopcoded marker.

Not part of saved state.

Prior to A.02.00

Next Pk Right

Pressing Next Pk Right moves the selected marker to the nearest peak right of the current marker which meets all enabled peak criteria. If there is no valid peak to the right of the current marker position, an error is generated and the marker is not moved.

If the selected marker was off, then it is turned on as a normal marker and a peak search is performed.

Key Path

Remote Command

Example

Notes

State Saved

Initial S/W Revision

Peak Search

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MAXimum:R

IGHt

CALC:MARK2:MAX:RIGH !Selects marker 2 and moves it to the next peak to the right of the current marker position.

Sending this command selects the subopcoded marker.

Not part of saved state.

Prior to A.02.00

Next Pk Left

Pressing Next Pk Left moves the selected marker to the nearest peak left of the current marker which meets all enabled peak criteria. If there is no valid peak to the left of the current marker position, an error

242

Common Measurement Functions 1

is generated and the marker is not moved.

If the selected marker was off, then it is turned on as a normal marker and a peak search is performed.

Key Path

Remote Command

Example

State Saved

Initial S/W Revision

Peak Search

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MAXimum:L

EFT

CALC:MARK2:MAX:LEFT selects marker 2 and moves it to the next peak to the left of the current marker position.

Not part of saved state.

Prior to A.02.00

Marker Delta

Performs the same function as the Delta 1-of-N selection key in the Marker menu. Basically this sets the control mode for the selected marker to Delta mode. See the Section “ “Marker” on page 1571 ” for the complete description of this function. The key is duplicated here in the Peak Search Menu to allow you to conveniently perform a peak search and change the marker’s control mode to Delta without having to access two separate menus.

Initial S/W Revision

Key Path

Notes

Prior to A.02.00

Peak Search or Marker

Whenever the selected marker is in Delta mode and you are in the Peak Search menu, the Marker Delta key should be highlighted and the active function for setting its delta value turned on.

Mkr->CF

Assigns the selected marker’s frequency to the Center Frequency setting. See the Section “ “Marker To” on page 1613 ” for the description of this function. The key is duplicated here in the Peak Search Menu to allow you to conveniently perform a peak search and marker to CF without having to access two separate menus.

Key Path

Dependencies

Initial S/W Revision

Peak Search or Marker ->

Same as specified under Marker To

Prior to A.02.00

Mkr->Ref Lvl

Assigns the selected marker’s level to the Reference Level setting. See the Section “ “Marker To” on page 1613 ” for the description of this function. The key is duplicated here in the Peak Search Menu to allow you to conveniently perform a peak search and marker to RL without having to access two separate menus.

Key Path

Peak Search or Marker ->

243

Common Measurement Functions 1

Dependencies

Initial S/W Revision

Same as specified under Marker To

Prior to A.02.00

Peak Criteria

Pressing this key opens the Peak Criteria menu and allows you to adjust the Pk Threshold and Pk

Excursion parameters used for peak search functions.

For a signal to be identified as a peak it must meet certain criteria. Signals in the negative frequency range and signals very close to 0 Hz are ignored. If either the peak excursion or peak threshold functions are on, then the signal must satisfy those criteria before being identified as a peak.

When peak excursion and peak threshold are both off:

Peak Search, Continuous Peak Search,

and maximum part of

Pk-Pk Search

will search the trace for the point with the highest y-axis value which does not violate the LO feedthrough rules. A rising and falling slope are not required for these three peak search functions.

The remaining search functions

Next Peak, Next Pk Right

, etc. will only consider trace points which have a rising and falling slope on the left and right respectively.

Key Path

Initial S/W Revision

Peak Search

Prior to A.02.00

“Peak Search” Criteria

This menu lets you decide what kind of search you want to do when the Peak Search key is pressed (or the equivalent SCPI command sent).

Note that there are two “types” of peak search functions. One type is the “Peak Search” type, the other type is the “Next Peak” type. “Next Peak” searches (for example, Next Peak, Next Pk Left, Next Pk

Right) are always checked using the Excursion and Threshold criteria as long as these criteria are

On.The “Peak Search” type of search, simply finds the highest point on the trace. However you can change the “Peak Search” type of search so that it also uses the Excursion and Threshold criteria. This allows you to find the Maximum point on the trace that also obeys the Excursion and/or Threshold criteria.

When

Highest Peak

is selected, pressing

Peak Search

simply finds the highest peak on the marker’s trace. If

Same as “Next Peak” Criteria

is selected, then the search is also forced to consider the

Excursion and Threshold found under the “

Next Peak” Criteria

menu.

Key Path

Remote Command

Notes

Preset

Peak Search, Peak Criteria

:CALCulate:MARKer:PEAK:SEARch:MODE MAXimum|PARameter

:CALCulate:MARKer:PEAK:SEARch:MODE?

MAXimum corresponds to the

Highest Peak

setting

PARameter corresponds to the

Same as “Next Peak” Criteria

setting

MAXimum

244

Common Measurement Functions 1

State Saved

Readback line

Initial S/W Revision

Saved in instrument state.

Current state

Prior to A.02.00

Highest Peak

When this key is selected, pressing the Peak Search key or issuing the equivalent SCPI command finds the maximum point on the trace, subject to the peak-search qualifications. This also affects the Peak Search half of

Pk-Pk search and the Continuous Peak Search.

Key Path

Example

Readback

Initial S/W Revision

Peak Search, Peak Criteria, “Peak Search” Criteria

CALC:MARK:PEAK:SEAR:MODE MAX

Highest Peak

Prior to A.02.00

Same as “Next Peak” Criteria

When this key is selected, pressing the Peak Search key or issuing the equivalent SCPI command finds the maximum point on the trace, but subject to the Excursion and Threshold set under the Next Peak Criteria menu.

The search is, of course, also subject to the peak-search qualifications. This also affects the Peak Search half of

Pk-Pk search and the Continuous Peak Search.

Key Path

Example

Readback

Initial S/W Revision

Peak Search, Peak Criteria, “Peak Search” Criteria

CALC:MARK:PEAK:SEAR:MODE PAR

Use Excurs & Thr

Prior to A.02.00

“Next Peak” Criteria

This key opens up a menu which allows you to independently set the Peak Excursion and Peak

Threshold and turn them on and off.

Key Path

Initial S/W Revision

Peak Search, Peak Criteria

Prior to A.02.00

Pk Excursion On/Off

Turns the peak excursion requirement on/off and sets the excursion value. The value defines the minimum amplitude variation (rise and fall) required for a signal to be identified as peak. For example, if a value of

6 dB is selected, peak search functions like the marker Next Pk Right function move only to peaks that rise and fall

6 dB or more.

When both Pk Excursion and Pk Threshold are on, a signal must rise above the Pk Threshold value by at least the

Peak Excursion

value and then fall back from its local maximum by at least the

Peak Excursion

value to be considered a peak.

245

Common Measurement Functions 1

NOTE

In the event that a sequence of trace points with precisely the same values represents the maximum, the leftmost point is found.

See “More Information” on page 246 .

Key Path

Remote Command

Example

Dependencies

Couplings

Preset

Preset

State Saved

Min

Max

Initial S/W Revision

Peak Search, Peak Criteria, “Next Peak” Criteria

:CALCulate:MARKer:PEAK:EXCursion <rel_ampl>

:CALCulate:MARKer:PEAK:EXCursion?

:CALCulate:MARKer:PEAK:EXCursion:STATe OFF|ON|0|1

:CALCulate:MARKer:PEAK:EXCursion:STATe?

:CALC:MARK:PEAK:EXC:STAT ON

:CALC:MARK:PEAK:EXC 30 DB sets the minimum peak excursion requirement to 30 dB

Available only when Y axis unit is amplitude units, otherwise grayed out.

Whenever you adjust the value of Pk Excursion (with the knob, step keys, or by completing a numeric entry), and Peak Threshold is turned ON, the Peak

Threshold Line and the Peak Excursion Region are displayed.

6.0 dB

ON

6.0 dB

ON

Saved in State

0.0 dB

100.0 dB

Prior to A.02.00

More Information

If two signals are very close together and the peak excursion and threshold criteria are met at the outside edges of the combined signals, this function finds the highest of these two signals as a peak (or next peak). However, if a signal appears near the edge of the screen such that the full extent of either the rising or falling edge cannot be determined, and the portion that is on screen does not meet the excursion criteria, then the signal cannot be identified as a peak.

When measuring signals near the noise floor, you can reduce the excursion value even further to make these signals recognizable. To prevent the marker from identifying noise as signals, reduce the noise floor variations to a value less than the peak-excursion value by reducing the video bandwidth or by using trace averaging.

Pk Threshold On/Off

Turns the peak threshold requirement on/off and sets the threshold value. The peak threshold value defines the minimum signal level (or min threshold) that the peak identification algorithm uses to recognize a peak.

246

Common Measurement Functions 1

When both Pk Excursion and Pk Threshold are on, a signal must rise above the Pk Threshold value by at least the

Peak Excursion

value and then fall back from its local maximum by at least the

Peak Excursion

value to be considered a peak.

For example, if a threshold value of –90 dBm is selected, the peak search algorithm will only consider signals with amplitude greater than the –90 dBm threshold. If a threshold value of –90 dBm is selected, and

Peak Excursion

is

On

and set to 6 dB, the peak search algorithm will only consider signals with amplitude greater than the –90 dBm threshold which rise 6 dB above the threshold and then fall back to the threshold.

Key Path

Remote Command

Example

Dependencies

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

Peak Search, Peak Criteria, “Next Peak Criteria”

:CALCulate:MARKer:PEAK:THReshold <ampl>

:CALCulate:MARKer:PEAK:THReshold?

:CALCulate:MARKer:PEAK:THReshold:STATe OFF|ON|0|1

:CALCulate:MARKer:PEAK:THReshold:STATe?

CALC:MARK:PEAK:THR:STAT ON turns on the threshold criterion.

CALC:MARK:PEAK:THR –60 dBm sets the threshold to –60 dBm.

When Ref Level Offset changes, Peak Threshold must change by the same amount.

–90.0 dBm

ON

Saved in instrument state.

The current displayed Ref Level – 200 dB. The current displayed Ref Level is the current Ref Level, offset by the Ref Level Offset.

The current displayed Ref Level. This means the current Ref Level, offset by the Ref Level Offset.

depends on the current selected Y axis unit

Prior to A.02.00

Pk Threshold Line On/Off

Turns the peak threshold line on or off. Preset state is off. No equivalent SCPI command.

See

“More Information” on page 247

.

Dependencies

Initial S/W Revision

If Peak Threshold is Off and the Peak Threshold line is turned on, it should turn on Peak Threshold.

Prior to A.02.00

More Information

The Peak Threshold line is green and has the value of the peak threshold (for example, “–20.3 dBm”) written above its right side, above the line itself. If Peak Excursion is ON it shows on the left side as a region above the Peak

Threshold line. As with all such lines (Display Line, Trigger Level line, etc) it is drawn on top of all traces.

247

Common Measurement Functions 1

This function is automatically set to ON (thus turning on the Peak Threshold line) whenever the value of Peak

Threshold or Peak Excursion becomes the active function, unless Peak Threshold is OFF. It is automatically set to

OFF whenever Peak Threshold is set to OFF. Manually turning it ON automatically turns on Pk Threshold.

The Peak Excursion part is on whenever the Pk Threshold part is on, unless Peak Excursion is OFF.

Peak Table

Opens the Peak Table menu.

The Peak Table provides a displayed list of up to 20 signal peaks from the selected trace. If more than one trace window is displayed, the selected trace in the selected window is used. If there are more than

20 signals which meet the peak search criteria, only the 20 highest peaks are listed.

The Peak Table is updated after each sweep. The list of peaks in the Peak Table can be ordered either by ascending frequency or by descending amplitude. In either case, the entire trace is first evaluated and the

20 highest peaks are selected for inclusion in the list. After the peaks are selected, they are then sorted and displayed according to the Peak Sort setting.

Key Path

Initial S/W Revision

Peak Search

Prior to A.02.00

Peak Table On/Off

Turns Peak Table on/off. When turned on, the display is split into a measurement window and a peak table display window.

248

Common Measurement Functions 1

Turning the Peak Table on turns the Marker Table off and vice versa.

Key Path

Remote Command

Example

Dependencies

Preset

State Saved

Initial S/W Revision

Peak Search, Peak Table

:CALCulate:MARKer:PEAK:TABLe:STATe OFF|ON|0|1

:CALCulate:MARKer:PEAK:TABLe:STATe?

CALC:MARK:PEAK:TABL:STAT ON Turns on and displays the peak table.

When the Peak Table turns on, if Peak Threshold is On then it becomes the active function.

OFF

Saved in instrument state.

Prior to A.02.00

Peak Sort

Sets the peak table sorting routine to list the peaks in order of descending amplitude or ascending frequency. The remote command can also be used to sort the peaks found using the

:CALCulate:DATA:PEAKs command.

Key Path

Remote Command

Example

Preset

Preset

State Saved

Backwards Compatibility SCPI

Backwards Compatibility SCPI

Initial S/W Revision

Peak Search, Peak Table

:CALCulate:MARKer:PEAK:SORT FREQuency|AMPLitude

:CALCulate:MARKer:PEAK:SORT?

CALC:MARK:PEAK:SORT AMPL Sets sorting routine to list peaks in order of descending amplitude.

CALC:MARK:PEAK:SORT?

AMPLitude

AMPLitude

Saved in instrument state.

:TRACe:MATH:PEAK:SORT

The old TRAC:MATH:PEAK:SORT command/query used in ESA is still supported for backward compatibility.

Prior to A.02.00

Peak Readout

Shows up to twenty signal peaks as defined by the setting:

All (ALL) - lists all the peaks defined by the peak criteria, in the current sort setting.

Above Display Line (GTDLine) - lists the peaks that are greater than the defined display line, and that meet the peak criteria. They are listed in the current sort order.

Below Display Line (LTDLine) - lists the peaks that are less than the defined display line, and that meet

249

Common Measurement Functions 1

the peak criteria. They are listed in the current sort order.

If the peak threshold is defined and turned on, then the peaks must meet this peak criteria in addition to the display line requirements.

See

“More Information” on page 250 .

Key Path

Remote Command

Example

Dependencies

Preset

Preset

State Saved

Readback line

Initial S/W Revision

Peak Search, Peak Table

:CALCulate:MARKer:PEAK:TABLe:READout

ALL|GTDLine|LTDLine

:CALCulate:MARKer:PEAK:TABLe:READout?

CALC:MARK:PEAK:TABL:READ GTDL

Turning Display Line off forces Readout to ALL

All

All

Saved in instrument state.

1-of-N selection

Prior to A.02.00

More Information

If the Display Line (see the Section “View/Display”) is turned on, the Peak Table can be selected to include all peaks, only those above the Display Line, or only those below the Display Line. See Figures

1–2 and 1–3 to understand what happens if both Display Line and Pk Threshold are turned on.

peak listed peak not listed peak listed peak not listed

Display Line

Threshold

Threshold

Display Line

pk_abovedl.vsd

Figure 1- 2Above Display Line Peak Identification

250

Common Measurement Functions 1

peak not listed peak listed

Display Line

Threshold

peak not listed

Threshold

Display Line

pk_belowdl.vsd

Figure 1- 3Below Display Line Peak Identification

All

Sets the peak table to display the 20 highest peaks in the order specified by the current Peak Sort setting. If the Peak

Criteria are turned on, then only peaks that meet the defined Pk Excursion and Pk Threshold values will be found.

Key Path

Example

Notes

Readback

Initial S/W Revision

Peak Search, Peak Table, Peak Readout

CALC:MARK:PEAK:TABL:READ ALL

Auto return after pressed

All

Prior to A.02.00

Above Display Line

Sets the peak table to display only the 20 highest peaks above the display line in the order specified by the current

Sort setting. If the Peak Criteria are turned on, then only peaks that meet the defined criteria will be found. If the display line is not already on, it is turned on (it has to be on or it cannot be used to exclude peaks).

Key Path

Example

Notes

Dependencies

Peak Search, Peak Table, Peak Readout

CALC:MARK:PEAK:TABL:READ GTDL

Auto return after pressed

When Above Display Line is selected, Display Line is turned on and becomes the active function.

Above DL

Prior to A.02.00

Readback

Initial S/W Revision

Below Display Line

Sets the peak table to display only the 20 highest peaks below the display line as defined by the peak in the order specified by the current Sort setting. If the Peak Criteria are turned on, then only peaks that meet the defined criteria will be found. If the display line is not already on, it is turned on (it has to be on or it cannot be used to

251

Common Measurement Functions 1

exclude peaks).

Key Path

Example

Notes

Dependencies

Readback

Initial S/W Revision

Peak Search, Peak Table, Peak Readout

CALC:MARK:PEAK:TABL:READ LTDL

Auto return after pressed

When Below Display Line is selected, Display Line is turned on and becomes the active function.

Below DL

Prior to A.02.00

Continuous Peak Search On/Off

Turns Continuous Peak Search on or off. When Continuous Peak Search is on, a peak search is automatically performed for the selected marker after each sweep. The rules for finding the peak are exactly the same as for

Peak Search

, including the use of the peak criteria rules. If no valid peak is found, a warning is generated after each sweep.

See

“More Information” on page 252 .

Key Path

Remote Command

Example

Notes

Couplings

Preset

State Saved

Status Bits/OPC dependencies

Initial S/W Revision

Peak Search

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:CPSearch[

:STATe] ON|OFF|1|0

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:CPSearch[

:STATe]?

CALC:MARK:CPS ON Turns on Continuous Peak Search.

Sending this command selects the subopcoded marker

The Continuous Peak Search key is grayed out when the selected marker is a

Fixed

marker. Also, if Continuous Peak Search is on and the selected marker becomes a fixed marker, then Continuous Peak Search is turned off and the key grayed out.

Signal Track

and

Continuous Peak Search

are mutually exclusive so if

Signal Track

is on,

Continuous Peak Search

will be grayed out and vice versa.

Mode Preset

Saved in instrument state.

The Measuring bit should remain set while this command is operating and should not go false until the marker position has been updated.

Prior to A.02.00

More Information

When Continuous Peak Search is turned on a peak search is immediately performed and then is repeated

252

Common Measurement Functions 1

after each sweep. If Continuous Peak Search is turned on with the selected marker off, the selected marker is set to

Normal

at the center of the screen, and then a peak search is immediately performed and subsequently repeated after each sweep.

When in Continuous Peak Search, *OPC will not return true, nor will READ or MEASure return any data, until the sweep is complete and the marker has been re-peaked. Note further that if the analyzer is in a measurement such as averaging, and Continuous Peak Search is on, the entire measurement will be allowed to complete (i.e., all the averages taken up to the average number) before the repeak takes place, and only THEN will *OPC go true and READ or MEASure return data.

Note that this function is not the “Continuous Peak” function found in some other instruments. That function was designed to track the signal; this function simply does a Peak Search after each sweep.

When Continuous Peak Search is turned on for a marker, a little “hat” is placed above the marker.

Pk-Pk Search

Finds and displays the amplitude and frequency (or time, if in zero span) differences between the highest and lowest y-axis value. It places the selected marker on the minimum value on its selected trace. And it places that marker’s reference marker on the peak of its selected trace. This function turns on the reference marker and sets its mode to

Fixed

if it is not already on. (These markers may be on two different traces.)

The rules for finding the maximum peak are exactly the same as for

Peak Search

, including the use of the peak criteria rules. However, the minimum trace value is not required to meet any criteria other than being the minimum y-axis value in the trace.

When Pk-Pk Search is successful, a message is displayed on the message line.

If the selected marker is off, a delta type marker is turned on and the peak-to-peak search is done. If the selected marker is on, but it is not a delta marker, then it is changed to delta which turns on the reference marker if needed, and then it performs the peak-to-peak function.

Key Path

Remote Command

Example

Notes

Notes

Dependencies

Couplings

State Saved

Initial S/W Revision

Peak Search

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:PTPeak

CALC:MARK:PTP

CALC:MARK:Y? !queries the delta amplitude value for marker 1.

Turns on the Marker

Δ active function.

Sending this command selects the subopcoded marker.

Pk-Pk Search is grayed out when

Coupled Markers

is on.

The selected marker becomes a delta marker if not already in delta mode.

Not part of saved state.

Prior to A.02.00

Min Search

Moves the selected marker to the minimum y-axis value on the current trace. Minimum (negative) peak

253

Common Measurement Functions 1

searches do not have to meet the peak search criteria. It just looks for the lowest y-axis value. If the selected marker is Off, it is turned on before the minimum search is performed.

Key Path

Remote Command

Example

Notes

State Saved

Initial S/W Revision

Peak Search

:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MINimum

CALC:MARK:MIN selects marker 1 and moves it to the minimum amplitude value.

Sending this command selects the subopcoded marker.

Not part of saved state.

Prior to A.02.00

Peak Data Query (RemoteCommand Only)

This command works the same way in this and many other measurements, so it is documented in the

Common Measurement Functions section.. For details about this key, see

“Calculate Peaks of Trace

Data (Remote Command Only)” on page 164

.

254

Common Measurement Functions 1

Recall

Most of the functions under this key work the same way in many measurements, so they are documented

in the Utility Functions section. For details about this key, see “Recall” on page 139

.

The Amplitude Correction Import Data function under Recall is documented here.

Amplitude Correction

This key selects the Amplitude Corrections as the data type to be imported. When pressed a second time, it brings up the Select Menu, which lets you select the Correction into which the data will be imported.

Amplitude Corrections are fully discussed in the documentation of the Input/Output key, under the

Corrections softkey.

A set of preloaded Corrections files can be found in the directory

/My Documents/ EMC Limits and Ampcor.

Under this directory, the directory called Ampcor (Legacy Naming) contains a set of legacy corrections files, generally the same files that were supplied with older Agilent EMI analyzers, that use the legacy suffixes .ant, .oth, .usr, and .cbl, and the old 8-character file names. In the directory called Ampcor, the same files can be found, with the same suffixes, but with longer, more descriptive filenames.

When the Amplitude Correction is an Antenna correction and the Antenna Unit in the file is not

None

, the Y Axis Unit setting will change to match the Antenna Unit in the file.

Remote Command

Example

Initial S/W Revision

Key Path

Mode

:MMEMory:LOAD:CORRection 1|2|3|4, <filename>

:MMEM:LOAD:CORR 2 "myAmpcor.csv" recalls the Amplitude Correction data from the file myAmpcor.csv in the current directory to the 2nd Amplitude

Correction table, and turns on Correction 2.

The default path is My Documents\amplitudeCorrections.

A.02.00

Recall, Data

SA|EDGEGSM

255

Common Measurement Functions 1

Dependencies

Couplings

Readback

Backwards Compatibility SCPI

Only the first correction array (Correction 1) supports antenna units. This means that a correction file with an Antenna Unit can only be loaded into the

Corrections 1 register. Consequently only for Correction 1 does the dropdown in the Recall dialog include.ant, and if an attempt is made to load a correction file into any other Correction register which DOES contain an antenna unit, a

Mass Storage error is generated.

Corrections are not supported by all Measurements. If in a Mode in which some Measurements support it, this key will be grayed out in measurements that do not. The key will not show at all if no measurements in the Mode support it.

Errors are reported if the file is empty or missing, or if the file type does not match, or if there is a mismatch between the file type and the destination data type. If any of these occur during manual operation, the analyzer returns to the Import Data menu and the File Open dialog goes away.

This key does not appear unless you have the proper option installed in your instrument.

This command will generate an “Option not available” error unless you have the proper option installed in your instrument.

When a correction file is loaded from mass storage, it is automatically turned on (Correction ON) and Apply Corrections is set to On. This allows the user to see its effect, thus confirming the load. selected Correction

For backwards compatibility, the following parameters syntax is supported:

:MMEMory:LOAD:CORRection ANTenna|CABLe|OTHer|USER,

<filename>

ANTenna maps to 1, CABle maps to 2, OTHer maps to 3 and USER maps to 4

Amplitude Correction 1, 2, 3, 4

These keys let you select which Correction to import the data into. Once selected, the key returns back to the Import Data menu and the selected Correction number is annotated on the key. The next step is to select the Open key in the Import Data menu.

Antenna corrections are a particular kind of Amplitude Corrections – they are distinguished in the corrections file by having the Antenna Unit set to a value other than None. Only Correction 1 supports

Antenna Units.

Initial S/W Revision

Key Path

Notes

Dependencies

Preset

A.02.00

Recall, Data, Amplitude Correction

auto return

Only Correction 1 may be used to load a Correction that contains an

Antenna Unit other than None not part of Preset, but is reset to Correction 1 by Restore Input/Output

Defaults; survives shutdown

256

State Saved

Common Measurement Functions 1

The current Correction number is saved in State

257

Common Measurement Functions 1

Restart

The Restart function restarts the current sweep, or measurement, or set of averaged/held sweeps or measurements. If you are Paused, pressing Restart does a Resume.

The Restart function is accessed in several ways:

• Pressing the Restart key

• Sending the remote command INIT:IMMediate

• Sending the remote command INIT:RESTart

Remote Command:

Example:

Couplings:

:INITiate[:IMMediate]

:INIT:IMM

Resets average/hold count k. For the first sweep overwrites all active

(update=on) traces with new current data. For application modes, it resets other parameters as required by the measurement.

Notes:

Key Path:

Status Bits/OPC dependencies:

:INITiate:RESTart

:INITiate:IMMediate

Either of the above commands perform exactly the same function.

Front-panel key

This is an Overlapped command.

The STATus:OPERation register bits 0 through 8 are cleared.

The STATus:QUEStionable register bit 9 (INTegrity sum) is cleared.

The SWEEPING bit is set.

The MEASURING bit is set.

Initial S/W Revision: Prior to A.02.00

Remote Command:

Example:

Couplings:

Notes:

:INITiate:RESTart

:INIT:REST

Resets average/hold count k. For the first sweep overwrites all active

(update=on) traces with new current data. For application modes, it resets other parameters as required by the measurement.

:INITiate:RESTart

:INITiate:IMMediate

Either of the above commands perform exactly the same function.

258

Common Measurement Functions 1

Status Bits/OPC dependencies: This is an Overlapped command.

The STATus:OPERation register bits 0 through 8 are cleared.

The STATus:QUEStionable register bit 9 (INTegrity sum) is cleared.

The SWEEPING bit is set.

The MEASURING bit is set.

Initial S/W Revision: Prior to A.02.00

The Restart function first aborts the current sweep/measurement as quickly as possible. It then resets the sweep and trigger systems, sets up the measurement and initiates a new data measurement sequence with a new data acquisition (sweep) taken once the trigger condition is met.

If the analyzer is in the process of aligning when Restart is executed, the alignment finishes before the restart function is performed.

Even when set for Single operation, multiple sweeps may be taken when Restart is pressed (for example, when averaging/holding is on). Thus when we say that Restart "restarts a measurement," we may mean:

• It restarts the current sweep

• It restarts the current measurement

• It restarts the current set of sweeps if any trace is in Trace Average, Max Hold or Min Hold

• It restarts the current set of measurements if Averaging, or Max Hold, or Min Hold is on for the measurement

• depending on the current settings.

With Average/Hold Number (in Meas Setup menu) set to 1, or Averaging off, or no trace in Trace

Average or Hold, a single sweep is equivalent to a single measurement. A single sweep is taken after the trigger condition is met; and the analyzer stops sweeping once that sweep has completed. However, with

Average/Hold Number >1 and at least one trace set to Trace Average, Max Hold, or Min Hold (SA

Measurement) or Averaging on (most other measurements), multiple sweeps/data acquisitions are

taken for a single measurement. The trigger condition must be met prior to each sweep. The sweep is stopped when the average count k equals the number N set for Average/Hold Number. A measurement average usually applies to all traces, marker results, and numeric results; but sometimes it only applies to the numeric results.

Once the full set of sweeps has been taken, the analyzer will go to idle state. To take one more sweep without resetting the average count, increment the average count by 1, by pressing the step up key while

Average/Hold Number is the active function, or sending the remote command CALC:AVER:TCON

UP.

259

Common Measurement Functions 1

Save

Most of the functions under this key work the same way in many measurements, so they are documented in the Utility Functions section. For details about this key, see

“Save” on page 155 .

The Amplitude Correction Export Data function under Save is documented here.

Amplitude Correction

Pressing this key selects Amplitude Corrections as the data type to be exported. Pressing this key again brings up the Select Menu, which allows the user to select which Amplitude Correction to save.

Amplitude Corrections are fully discussed in the documentation of the Input/Output key, under the

Corrections softkey.

Remote Command:

Example:

Notes:

:MMEMory:STORe:CORRection 1|2|3|4, <filename>

:MMEM:STOR:CORR 2 "myAmpcor.csv" saves Correction 2 to the file myAmpcor.csv on the current path.

The default path is My Documents\amplitudeCorrections.

If the save is initiated via SCPI, and the file already exists, the file will be overwritten.

Using the C: drive is strongly discouraged, since it runs the risk of being overwritten during an instrument software upgrade.

Both single and double quotes are supported for any filename parameter over

SCPI.

Key Path:

Readback:

Save, Data

1 | 2 | 3 | 4

Dependencies: Corrections are not supported by all Measurements. If in a Mode in which some Measurements support it, this key will be grayed out in measurements that do not. The key will not show at all if no measurements in the Mode support it.

This key will not appear unless you have the proper option installed in your instrument.

Backwards Compatibility SCPI: For backwards compatibility only, the following parameters syntax is supported:

:MMEMory:STORe:CORRection ANTenna|CABLe|OTHer|USER,

<filename>

ANTenna maps to 1, CABle maps to 2, OTHer maps to 3 and USER maps to 4

Initial S/W Revision: A.02.00

Correction Data File

A Corrections Data File contains a copy of one of the analyzer correction tables. Corrections provide a

260

Common Measurement Functions 1

way to adjust the trace display for predetermined gain curves (such as for cable loss).

The first five lines are system-required header lines, and must be in the correct order.

Amplitude CorrectionData file type name

“Correction Factors for 11966E”File Description

“Class B Radiated”Comment

A.01.00.R0001,N9020AInstrument Version, Model Number

P13 EA3 UK6,01Option List, File Format Version

Corrections files may include Antenna amplitude units. This amplitude unit in the Antenna Unit field is a conversion factor that is used to adjust the Y Axis Units of the current mode, if the mode supports

Antenna Units. For more details on antenna correction data, refer to the Input/Output chapter,

Corrections section.

The metadata required to properly import the correction data is:

Frequency Unit for the x axis data

Antenna Unit for the y axis data (not required)

Frequency Interpolation algorithm – either Logarithmic or Linear

The data follows as comma separated X, Y pairs; one pair per line. The keyword “DATA” precedes the data.

For example, suppose you have an Antenna to correct for on an E4445A version A.01.00 R0011 and the correction data is:

0 dB at 200 MHz

17 dB at 210 MHz

14.8 dB at 225 MHz

Then the file will look like:

Amplitude Correction

"Correction Factors for 11966E"

"Class B Radiated"

A.01.00 R0011,N9020A

P13 EA3 UK6,01

Frequency Unit,MHz

Antenna Unit,dBuV/m

Frequency Interpolation,Linear

DATA

200.000000,0.00

210.000000,17.00

225.000000,14.80

261

Common Measurement Functions 1

The choices for the 1 of N fields in the metadata are as follows:

Frequency Unit: Hz, kHz, MHz, GHz

Antenna Unit: dBuv/m, dBuA/m, dBG, dBpT, None

Frequency Interpolation: Logarithmic, Linear

Amplitude Correction 1, 2, 3, 4

These keys let you pick which Correction to save. Once selected, the key returns back to the Export Data menu and the selected Correction number is annotated on the key.

The next step in the Save process is to select the Save As key in the Export Data menu.

Key Path

Preset

Readback

Initial S/W Revision

Save, Data, Amplitude Correction

Not part of a Preset, but is reset to Correction 1 by Restore Input/Output

Defaults. Survives a shutdown.

1

A.02.00

262

Common Measurement Functions 1

Single (Single Measurement/Sweep)

Sets the analyzer for Single measurement operation. The single/continuous state is Meas Global, so the setting will affect all the measurements. If you are Paused, pressing

Single

does a Resume.

Example:

Notes:

Key Path:

Initial S/W Revision:

:INIT:CONT OFF

See

Cont

key description.

Front-panel key

Prior to A.02.00

Source

This mode does not have any Source control functionality.

Key Path

Initial S/W Revision

Front-panel key

Prior to A.02.00

Sweep/Control

Accesses a menu that enables you to configure the Sweep and Control functions of the analyzer, such as

Sweep Time and Gating.

Key Path

Initial S/W Revision

Front-panel key

Prior to A.02.00

Sweep Time

Controls the time the analyzer takes to sweep the current frequency span when the Sweep Type is Swept, and displays the equivalent Sweep Time when the Sweep Type is FFT.

When Sweep Time is in Auto, the analyzer computes a sweep time which will give accurate measurements based on other settings of the analyzer, such as RBW and VBW.

NOTE

The Meas Uncal (measurement uncalibrated) warning is given in the Status Bar in the lower right corner of the screen when the manual sweep time entered is faster than the sweep time computed by the analyzer’s sweep time equations, that is, the

Auto Sweep Time. The analyzer’s computed sweep time will give accurate measurements; if you sweep faster than this your measurements may be inaccurate. A Meas Uncal condition may be corrected by returning the Sweep

Time to Auto; by entering a longer Sweep Time; or by choosing a wider RBW and/or VBW

On occasion other factors such as the YTF sweep rate (in high band) or the LO’s

263

Common Measurement Functions 1

capability (in low band) can cause a Meas Uncal condition. The most reliable way to correct it is to return the Sweep Time to Auto.

When Sweep Type is FFT, you cannot control the sweep time, it is simply reported by the analyzer to give you an idea of how long the measurement is taking.

Note that although some overhead time is required by the analyzer to complete a sweep cycle, the sweep time reported when Sweep Type is Swept does not include the overhead time, just the time to sweep the

LO over the current Span. When Sweep Type is FFT, however, the reported Sweep Time takes into account both the data acquisition time and the processing time, in order to report an equivalent Sweep

Time for a meaningful comparison to the Swept case.

Because there is no “Auto Sweep Time” when in zero span, the Auto/Man line on this key disappears when in Zero Span. The Auto/Man line also disappears when in an FFT sweep. In this case the key is grayed out as shown below.

Key Path

Remote Command

Example

Notes

Dependencies

Sweep/Control

[:SENSe]:SWEep:TIME <time>

[:SENSe]:SWEep:TIME?

[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1

[:SENSe]:SWEep:TIME:AUTO?

SWE:TIME 500 ms

SWE:TIME:AUTO OFF

The values shown in this table reflect the “swept spans” conditions which are the default settings after a preset. See “Couplings” for values in the zero span domain.

The third line of the softkey (Auto/Man) disappears in Zero Span. The SCPI command SWEep:TIME:AUTO ON if sent in Zero Span generates an error message.

Softkey grayed out and third line of the softkey (Auto/Man) disappears in FFT sweeps. Pressing the key or sending the SCPI for sweep time while the instrument is in FFT sweep generates a –221, “Settings Conflict;” error. The

SCPI command :SWEep:TIME:AUTO ON if sent in FFT sweeps generates an error.

Grayed out while in Gate View, to avoid confusing those who want to set Gate

View Sweep Time.

Key is grayed out in Measurements that do not support swept mode.

Key is blanked in Modes that do not support swept mode.

Set to Auto when Auto Couple is pressed or sent remotely

264

Common Measurement Functions 1

Couplings

Preset

State Saved

Min

Max

Status Bits/OPC dependencies

Sweep Time is coupled primarily to Span and RBW. Center Frequency, VBW, and the number of sweep points also can have an effect. So changing these parameters may change the sweep time.

The Sweep Time used upon entry to Zero Span is the same as the Sweep Time that was in effect before entering Zero Span. The Sweep Time can be changed while in Zero Span. Upon leaving Zero Span, the Auto/Man state of Sweep

Time that existed before entering Zero Span is restored.

If Sweep Time was in Auto before entering Zero Span, or if it is set to Auto while in zero span (which can happen via remote command or if

Auto Couple

is pressed) it returns to Auto and recouples when returning to non-zero spans.

If Sweep Time was in Man before entering Zero Span, it returns to Man when returning to non-zero spans, and any changes to Sweep Time that were made while in Zero Span are retained in the non-zero span (except where constrained by minimum limits, which are different in and out of zero span).

The preset Sweep Time value is hardware dependent since Sweep Time presets to “Auto”.

Saved in instrument state in zero span: 1

μs in swept spans: 1 ms in zero span: 6000 s in swept spans: 4000 s

Meas Uncal is Bit 0 in the STATus:QUEStionable:INTegrity:UNCalibrated register

Prior to A.02.00

Initial S/W Revision

Sweep Setup

Lets you set the sweep functions that control features such as sweep type and time.

Key Path

Dependencies

Sweep/Control

The whole Sweep Setup menu is grayed out in Zero Span, however, the settings in the menus under Sweep Setup can be changed remotely with no error indication.

Grayed out in measurements that do not support swept mode.

Blanked in modes that do not support swept mode

Prior to A.02.00

Initial S/W Revision

Sweep Time Rules

Allows the choice of three distinct sets of sweep time rules. These are the rules that are used to set the sweep time when

Sweep Time

is in

Auto mode.

Note that these rules only apply when in the Swept

Sweep Type

(either manually or automatically chosen) and not when in FFT sweeps.

265

Common Measurement Functions 1

See

“More Information” on page 266 .

Key Path

Remote Command

Example

Dependencies

Couplings

Preset

State Saved

Backwards Compatibility SCPI

Initial S/W Revision

Sweep/Control, Sweep Setup

[:SENSe]:SWEep:TIME:AUTO:RULes

NORMal|ACCuracy|SRESponse

[:SENSe]:SWEep:TIME:AUTO:RULes?

SWE:TIME:AUTO:RUL ACC

In Zero Span, this key is irrelevant and cannot be accessed (because the whole

Sweep Setup menu is grayed out in Zero Span), however its settings can be changed remotely with no error indication.

Grayed out in FFT sweeps. Pressing the key while the instrument is in FFT sweep generates an advisory message. The SCPI is acted upon if sent, but has no effect other than to change the readout on the key, as long as the analyzer is in an FFT sweep.

Set to Auto on Auto Couple

AUTO

Saved in instrument state

The old Auto Sweep Time command was the same

[:SENSe]:SWEep:TIME:AUTO:RULes NORMal|ACCuracy so it still works although it now has a third parameter (SRESponse).

The old Sweep Coupling command was

[:SENSe]:SWEep:TIME:AUTO:MODE SRESponse|SANalyzer and it is aliased as follows:

:SWEep:TIME:AUTO:MODE SRESponse is aliased to

:SWEep:TIME:AUTO:RULes SRESponse, and

:SWEep:TIME:AUTO:MODE SANalyzer is aliased to

:SWEep:TIME:AUTO:RULes NORMal

The query :SWEep:TIME:AUTO:MODE? is aliased to

:SWEep:TIME:RULes?

So it will fail to match for SANalyzer

Prior to A.02.00

More Information

The first set of rules is called

SA – Normal

.

Sweep Time Rules

is set to

SA-Normal

on a

Preset

or

Auto Couple

. These rules give optimal sweep times at a lossof accuracy. Note that this means that in the

Preset or Auto Coupled state, instrument amplitude accuracy specifications do not apply.

Setting

Sweep Time Rules

to

SA-Accuracy

will result in slower sweep times than

SA-Normal

, usually about three times as long, but with better amplitude accuracy for CW signals. The instrument absolute amplitude accuracy specifications only apply when

Sweep Time

is set to

Auto

, and

Sweep Time Rules

266

Common Measurement Functions 1

are set to

SA-Accuracy

. Additional amplitude errors which occur when

Sweep Time Rules

are set to

SA-Normal

are usually well under 0.1 dB, though this is not guaranteed. Because of the faster sweep times and still low errors,

SA-Normal

is the preferred setting of

Sweep Time Rules

.

The third set of sweep time rules is called

Stimulus/Response

and is automatically selected when an integrated source is turned on, such as a Tracking Generator or a synchronized external source. The sweep times for this set of rules are usually much faster for swept-response measurements.

Stimulus-response auto-coupled sweep times are typically valid in stimulus-response measurements when the system’s frequency span is less than 20 times the bandwidth of the device under test. You can select these rules manually (even if not making Stimulus-Response measurements) to get faster sweeps without the “Meas Uncal” warning, but you are then not protected from the over-sweep condition and may end up with uncalibrated results. However, it is commonplace in measuring non-CW signals such as noise to be able to get excellent measurement accuracy at sweep rates higher than those required for CW signal accuracy, so this is a valid measurement technique.

Auto

Sets the analyzer to automatically choose the Sweep Time Rules for the measurement.

Key Path

Remote Command

Example

Couplings

Preset

Initial S/W Revision

Sweep/Control, Sweep Setup, Sweep Time Rules

[:SENSe]:SWEep:TIME:AUTO:RULes:AUTO[:STATe] ON|OFF|1|0

[:SENSe]:SWEep:TIME:AUTO:RULes:AUTO[:STATe]?

:SWE:TIME:AUTO:RUL:AUTO ON

Set on Preset or Auto Couple

ON

Prior to A.02.00

SA - Normal

Chooses Sweep Time Auto Rules for optimal speed and generally sufficient accuracy.

Key Path

Example

Couplings

Readback

Initial S/W Revision

Sweep/Control, Sweep Setup, Sweep Time Rules

:SWE:TIME:AUTO:RUL NORM

Automatically selected unless Source is on

If directly selected, sets AUTO to Off

SA - Normal

Prior to A.02.00

SA - Accuracy

Chooses Sweep Time Auto Rules for specified absolute amplitude accuracy.

Key Path

Example

Sweep/Control, Sweep Setup, Sweep Time Rules

:SWE:TIME:AUTO:RUL ACC

267

Common Measurement Functions 1

Dependencies

Couplings

Readback

Initial S/W Revision

Do not allow sweep time to fall below 20 ms when in SA - Accuracy

If directly selected, sets AUTO to Off

SA - Accuracy

Prior to A.02.00

Stimulus/Response

Chooses Sweep Time Auto Rules for use with a source.

Key Path

Example

Couplings

Readback

Initial S/W Revision

Sweep/Control, Sweep Setup, Sweep Time Rules

:SWE:TIME:AUTO:RUL SRES

Automatically selected when Source is on

If directly selected sets AUTO to Off

SR

Prior to A.02.00

Sweep Type

Chooses between the FFT and Sweep types of sweep.

Sweep Type refers to whether or not the instrument is in Swept or FFT analysis. When in Auto, the selection of sweep type is governed by two different sets of rules, depending on whether you want to optimize for dynamic range or for speed.

FFT “sweeps” should not be used when making EMI measurements; therefore, when a CISPR detector

(Quasi Peak, EMI Average, RMS Average) is selected for any active trace (one for which Update is on), the FFT key in the Sweep Type menu is grayed out, and the Auto Rules only choose Swept. If Sweep

Type is manually selected to be FFT, the CISPR detectors are all grayed out.

FFT sweeps will never be auto-selected when Screen Video, Log Video or Linear Video are the selected

Analog Output.

Key Path

Remote Command

Notes

Sweep/Control, Sweep Setup

[:SENSe]:SWEep:TYPE FFT|SWEep

[:SENSe]:SWEep:TYPE?

For a backward compatibility, the following remote parameters

AUTO

SWP will be supported by the [:SENSe]:SWEep:TYPE command.

268

Common Measurement Functions 1

Dependencies

Preset

Backwards Compatibility SCPI

In Zero Span, this key is irrelevant and cannot be accessed (because the whole

Sweep Setup menu is grayed out in Zero Span), however its settings can be changed remotely with no error indication.

When Gate is on, Gate Method selection affects Sweep Type:

Method FFT&Sweep menu

FFT Swept grayed out and rules choose FFT

Video FFT grayed out and rules choose Swept

LO FFT grayed out and rules choose Swept

AUTO

[:SENSe]:SWEep:TYPE AUTO !sets sweep type Auto to On

[:SENSe]:SWEep:TYPE SWP !selects sweep type Swept

Prior to A.02.00

Initial S/W Revision

Auto

When in Auto, the selection of sweep type is governed by two different sets of rules, depending on whether you want to optimize for dynamic range or for speed. These rules are chosen under the

Sweep Type Rules

key.

Key Path

Remote Command

Example

Couplings

Preset

State Saved

Initial S/W Revision

Sweep/Control, Sweep Setup, Sweep Type

[:SENSe]:SWEep:TYPE:AUTO OFF|ON|0|1

[:SENSe]:SWEep:TYPE:AUTO?

:SWE:TYPE:AUTO ON

Pressing Auto Couple always sets Sweep Type to Auto.

ON

Saved in instrument state

Prior to A.02.00

Swept

Manually selects swept analysis, so it cannot change automatically to FFT.

Key Path

Example

Dependencies

Couplings

Sweep/Control, Sweep Setup, Sweep Type

SWE:TYPE SWE

Grayed out while in Gated FFT (meaning Gate is ON and Gate Method is

FFT).

If this key is selected, the gate method Gated FFT is grayed out.

This selection is chosen automatically if any of the CISPR detectors is chosen for any active trace, in which case the FFT Sweep Type selection is also grayed out.

269

Common Measurement Functions 1

State Saved

Readback

Initial S/W Revision

Modified at S/W Revision

Saved in instrument state

Swept

Prior to A.02.00

A.02.00

FFT

Manually selects FFT analysis, so it cannot change automatically to Swept.

Key Path

Example

Dependencies

Couplings

State Saved

Readback

Initial S/W Revision

Sweep/Control, Sweep Setup, Sweep Type

SWE:TYPE FFT

When a CISPR detector (Quasi Peak, EMI Average, RMS Average) is selected for any active trace, the FFT key is grayed out.

If Sweep Type is manually selected to be FFT, the CISPR detectors are all grayed out.

Grayed out while in Gated LO (meaning Gate is ON and Gate Method is LO).

Grayed out while in Gated Video (meaning Gate is ON and Gate Method is

Video).

If this key is selected, all of the gate Methods except Gated FFT are grayed out.

If Manual FFT is selected, the Signal ID key is grayed out.

When Signal ID is on, Manual FFT is grayed out.

For both the dynamic range case and the speed case, swept is chosen whenever any form of Signal ID is on.

Saved in instrument state

FFT

Prior to A.02.00

Sweep Type Rules

Selects which set of rules will be used for automatically choosing the Sweep Type when Sweep Type is in Auto.

Key Path

Remote Command

Dependencies

Preset

Sweep/Control, Sweep Setup

[:SENSe]:SWEep:TYPE:AUTO:RULes SPEed|DRANge

[:SENSe]:SWEep:TYPE:AUTO:RULes?

In Zero Span, this key is irrelevant and cannot be accessed (because the whole

Sweep Setup menu is grayed out in Zero Span), however its settings can be changed remotely with no error indication.

DRANge

270

Common Measurement Functions 1

State Saved

Initial S/W Revision

Saved in instrument state

Prior to A.02.00

Auto

This selection is automatically chosen when Auto Couple is pressed. When in Auto, the Sweep Type Rules are set to Best Dynamic Range. It seems like a very simple Auto function but the use of this construct allows a consistent statement about what the Auto Couple key does.

Key Path

Remote Command

Example

Couplings

Preset

State Saved

Initial S/W Revision

Sweep/Control, Sweep Setup, Sweep Type Rules

[:SENSe]:SWEep:TYPE:AUTO:RULes:AUTO[:STATe] OFF|ON|0|1

[:SENSe]:SWEep:TYPE:AUTO:RULes:AUTO[:STATe]?

:SWE:TYPE:AUTO:RUL:AUTO ON

Pressing Auto Couple always sets Sweep Type Rules to Auto.

ON

Saved in instrument state

Prior to A.02.00

Best Dynamic Range

This selection tells the analyzer to choose between swept and FFT analysis with the primary goal of optimizing dynamic range. If the dynamic range is very close between swept and FFT, then it chooses the faster one. This auto selection also depends on RBW Type.

In determining the Swept or FFT setting, the auto rules use the following approach:

If the RBW Filter Type is Gaussian use the RBW for the Normal Filter BW and if that RBW > 210 Hz, use swept; for RBW <= 210 Hz, use FFT

If the RBW Filter Type is Flat Top, use the same algorithm but use 420 Hz instead of 210 Hz for the transition point between Swept and FFT

If any of the CISPR detectors is chosen for any active trace, always use Swept.

Key Path

Example

Couplings

Readback

Initial S/W Revision

Sweep/Control, Sweep Setup, Sweep Type Rules

SWE:TYPE:AUTO:RUL DRAN sets the auto rules to dynamic range.

Directly selecting this setting sets AUTO to OFF.

Dynamic Range

Prior to A.02.00

Best Speed

This selection tells the analyzer to choose between FFT or swept analysis based on the fastest analyzer speed.

Key Path

Sweep/Control, Sweep Setup, Sweep Type Rules

271

Common Measurement Functions 1

Example

Couplings

Readback

Initial S/W Revision

SWE:TYPE:AUTO:RUL SPE sets the rules for the auto mode to speed

Directly selecting this setting sets AUTO to OFF.

Speed.

Prior to A.02.00

FFT Width

This menudisplays and controls the width of the FFT’s performed while in FFT mode. The “FFT width” is the range of frequencies being looked at by the FFT, sometimes referred to as the “chunk width” -- it is not the resolution bandwidth used when performing the FFT.

It is important to understand that this function does not directly set the FFT width, it sets the limit on the

FFT Width. The actual FFT width used is determined by several other factors including the Span you have set. Usually the instrument picks the optimal FFT Width based on the current setup; but on occasion you may wish to limit the FFT Width to be narrower than that which the instrument would have set.

NOTE

this function does not allow you to widen the FFT Width beyond that which the instrument might have set; it only allows you to narrow it. You might do this to improve the dynamic range of the measurement or eliminate nearby spurs from your measurement.

Note that the

FFT Width

setting will have no effect unless in an FFT sweep.

See

“More Information” on page 273

Remote Command

Example

Initial S/W Revision

Key Path

Notes

[:SENSe]:SWEep:FFT:WIDTh <real>

[:SENSe]:SWEep:FFT:WIDTh?

SWE:FFT:WIDT 167 kHz sets this function to “<167.4 kHz”

Prior to A.02.00

Sweep/Control, Sweep Setup

The parameter is in units of frequency.

For values sent from SCPI, the analyzer chooses the smallest value that is at least as great as the requested value.

Examples:

Parameter 3.99 kHz is sent over SCPI. Analyzer chooses 4.01 kHz

Parameter 4.02 kHz is sent over SCPI. Analyzer chooses 28.81 kHz

Parameter 8 MHz is sent over SCPI. Analyzer chooses 10 MHz

272

Common Measurement Functions 1

Dependencies

Couplings

Preset

State Saved

Min

Max

Backwards Compatibility SCPI

Modified at S/W Revision

In some models, the analog prefilters are not provided. In these models the

FFT Width

function is always in

Auto

. The FFT Width key is blanked in these models, and the SCPI commands are accepted without error but have no effect.

In Zero Span, this key is irrelevant and cannot be accessed (because the whole

Sweep Setup menu is grayed out in Zero Span). However, its settings can be changed remotely with no error indication.

The FFT Width affects the

ADC Dither

function (see Meas Setup key) and the point at which the instrument switches from Swept to FFT acquisition.

The Preset is Auto, but Preset will also pick Best Dynamic Range and hence this function will be set to ~Maximum

Saved in instrument state

4.01 kHz

The maximum available FFT width is dependent on the IF Bandwidth option.

For Option B10 the maximimum available width is 10 MHz; for Option B25 it is 25 MHz, and for Option B40 it is 40 MHz.

[:SENSe]:SWEep:FFT:SPAN:RATio <integer>

[:SENSe]:SWEep:FFT:SPAN:RATio?

The behavior of the analyzer when it receives this command is to compute the

“intended segment width” by dividing the Span by the FFTs/Span parameter, then converting this intended width to an actual width by using the largest available FFT Width that is still less than the intended segment width. The

“Span” used in this computation is whatever the Span is currently set to, whether a sweep has been taken at that Span or not.

A.04.00

Key Path

Remote Command

Example

Couplings

Preset

State Saved

Initial S/W Revision

Sweep/Control, Sweep Setup

[:SENSe]:SWEep:FFT:WIDTh:AUTO OFF|ON|0|1

[:SENSe]:SWEep:FFT:WIDTh:AUTO?

:SWE:FFT:WIDT:AUTO ON

Pressing Auto Couple always sets FFT Width to Auto.

ON

Saved in instrument state

Prior to A.02.00

More Information

An FFT measurement can only be performed over a limited span known as the “FFT segment”. Several segments may need to be combined to measure the entire span. For advanced FFT control in the

X-Series, you have direct control over the segment width using the

FFT Width

control. Generally, in automatic operation, the X-Series sets the segment width to be as wide as possible, as this results in the

273

Common Measurement Functions 1

fastest measurements.

However, in order to increase dynamic range, most X-series models provide a set of analog prefilters that precede the ADC. Unlike swept measurements, which pass the signal through a bandpass before the

ADC, FFT measurements present the full signal bandwidth to the ADC, making them more susceptible to overload, and requiring a lower signal level. The prefilters act to alleviate this phenomenon - they allow the signal level at the ADC to be higher while still avoiding an ADC overload, by eliminating signal power outside the bandwidth of interest, which in turn improves dynamic range.

Although narrowing the segment width can allow higher dynamic ranges some cases, this comes at the expense of losing some of the speed advantages of the FFT, because narrower segments require more acquisitions and proportionately more processing overhead.

However, the advantages of narrow segments can be significant. For example, in pulsed-RF measurements such as radar, it is often possible to make high dynamic range measurements with signal levels approaching the compression threshold of the analyzer in swept spans (well over 0 dBm), while resolving the spectral components to levels below the maximum IF drive level (about –8 dBm at the input mixer). But FFT processing experiences overloads at the maximum IF drive level even if the RBW is small enough that no single spectral component exceeds the maximum IF drive level. If you reduce the width of an FFT, an analog filter is placed before the ADC that is about 1.3 times as wide as the FFT segment width. This spreads out the pulsed RF in time and reduces the maximum signal level seen by the

ADC. Therefore, the input attenuation can be reduced and the dynamic range increased without overloading the ADC.

Further improvement in dynamic range is possible by changing the

FFT IF Gain

(in the

Meas Setup

menu of many measurements). If the segments are reduced in width,

FFT IF Gain

can be set to High, improving dynamic range.

Depending on what IF Bandwidth option you have ordered, there can be up to three different IF paths available in FFT sweeps, as seen in the diagram below:

~

10 MHz

10

MHz

ADC

~

25 MHz

25 MHz

~

40 MHz

40 MHz

ADC

The 10 MHz path is always used for Swept sweeps. It is always used for FFT sweeps as well, unless the user specifies ~25 MHz in which case the 25 MHz path will be used for FFT sweeps, or ~40 MHz, in

274

Common Measurement Functions 1

which case the 40 MHz path will be used for FFT sweeps. Note that, although each of these keys picks the specified path, the analyzer may choose an FFT width less than the full IF width, in order to optimize speed, trading off acquisition time versus processing time.

Pause/Resume

Pauses a measurement after the current data acquisition is complete.

When Paused, the label on the key changes to Resume. Pressing Resume un-pauses the measurement.

When you are Paused, pressing

Restart

,

Single

or

Cont

does a Resume.

Key Path

Remote Command

Dependencies

Sweep/Control

:INITiate:PAUSe

Grayed out in Measurements that do not support Pausing.

Blanked in Modes that do not support Pausing.

Initial S/W Revision Prior to A.02.00

Key Path

Remote Command

Dependencies

Sweep/Control

:INITiate:RESume

Grayed out in Measurements that do not support Pausing.

Blanked in Modes that do not support Pausing.

Initial S/W Revision Prior to A.02.00

Gate

Accesses a menu that enables you to control the gating function. The Gate functionality is used to view signals best viewed by qualifying them with other events.

Gate setup parameters are the same for all measurements – they do not change as you change measurements. Settings like these are called “Meas Global” and are unaffected by Meas Preset.

NOTE

Key Path

Scope

Readback

Initial S/W Revision

Sweep Time autocoupling rules and annotation are changed by Gate being on.

Sweep/Control

Meas Global

The state and method of Gate, as [Off, LO] or [On, Video]. Note that for measurements that only support gated LO, the method is nonetheless read back, but always as LO.

Prior to A.02.00

275

Common Measurement Functions 1

Gate On/Off

Turns the gate function on and off.

When the Gate Function is on, the selected Gate Method is used along with the gate settings and the signal at the gate source to control the sweep and video system with the gate signal. Not all measurements allow every type of Gate Methods.

When Gate is on, the annunciation in the measurement bar reflects that it is on and what method is used, as seen in the following "Gate: LO" annunciator graphic.

Key Path

Remote Command

Example

Dependencies

Couplings

Preset

State Saved

Range

Backwards Compatibility SCPI

Initial S/W Revision

Sweep/Control, Gate

[:SENSe]:SWEep:EGATe[:STATe] OFF|ON|0|1

[:SENSe]:SWEep:EGATe[:STATe]?

SWE:EGAT ON

SWE:EGAT?

The function is unavailable (grayed out) and Off when:

Gate Method is LO or Video and FFT Sweep Type is manually selected.

Gate Method is FFT and Swept Sweep Type is manually selected.

Marker Count is ON.

When Meas Method is RBW or FAST, this function is unavailable and the key is grayed out.

Whenever Gate is on, Meas Method, RBW or FAST is unavailable and keys for those are grayed out.

When Gate is on, Offset Res BW and Offset Video BW are ignored (if you set these values) and the measurement works as if all Offset Res BW and all

Offset Video BW are coupled with the Res BW and the Video BW under the

BW menu. When Gate is on, the Offset BW key in the Offset/Limit menu is grayed out.

Off

Saved in instrument state

On|Off

[:SENSe]:SWEep:TIME:GATE[:STATe]

Prior to A.02.00

Gate View On/Off

Turning on Gate View in the Swept SA measurement provides a single-window gate view display.

276

Common Measurement Functions 1

Turning on Gate View in other measurements shows the split-screen Gate View. In these measurements, when the Gate View is on, the regular view of the current measurement traces and results are reduced vertically to about 70% of the regular height. The Zero Span window, showing the positions of the Gate, is shown between the Measurement Bar and the reduced measurement window. By reducing the height of the measurement window, some of the annotation on the Data Display may not fit and is not shown.

Key Path

Remote Command

Example

Dependencies

Couplings

Preset

State Saved

Range

Initial S/W Revision

Sweep/Control, Gate

[:SENSe]:SWEep:EGATe:VIEW ON|OFF|1|0

[:SENSe]:SWEep:EGATe:VIEW?

SWE:EGAT:VIEW ON turns on the gate view.

In the Swept SA measurement:

In Gate View, the regular Sweep Time key is grayed out .

In the other measurements:

When you turn Gate View on, the lower window takes on the current state of the instrument. Upon leaving Gate View, the instrument takes on the state of the lower window.

When you turn Gate View on, the upper window Sweep Time is set to the gate view sweep time.

These couplings apply to the Swept SA measurement:

• When Gate View is turned on, the instrument is set to Zero Span.

• Gate View automatically turns off whenever a Span other than Zero is selected.

• Gate View automatically turns off ifyou pressthe Last Span key while in

Gate View, and the instrument returns to the Span it was in before entering

Gate View (even if that is Zero Span).

When Gate View is turned on, the sweep time used is the gate view sweep time. This is set according to the rules in

“Gate View Sweep Time” on page 279

.

• When Gate View is turned off, Sweep Time is set to the normal Swept SA measurement sweep time.

• If Gate View is on and Gate is off, then turning on Gate turns off Gate

View.

OFF

Saved in instrument state

On|Off

Prior to A.02.00

A sample of the Gate View screen in the Swept SA measurement is shown in the following graphic :

277

Common Measurement Functions 1

A sample of the Gate View screen in other measurements is shown in the following graphic . This example is for the ACP measurement:

278

Common Measurement Functions 1

Turning Gate View off returns the analyzer to the Normal measurement view.

In the Swept SA, the normal measurement view is the single-window Swept SA view. When returning to this view, the Swept SA measurement returns to the Span it was in before entering Gate View (even if that is Zero Span).

The Gate View window is triggered from the Gate Source, with zero trigger delay. Also, when updating the Gate View window, the Gate itself must not operate. So it is internally shut off while the gate view window is being updated. For the Swept SA measurement, this means that the Gate is internally shut off whenever the gate view window is displayed. The measurement bar and softkeys continue to show the

Trigger source for the main sweep window and give no indication that the Gate is shut off or that the

Gate View window is triggered from the Gate Source.

When in Gate View, vertical lines are displayed in the Gate View window as follows:

• Green lines are displayed at the gate edges as follows: in Edge Gate, a line is shown for Delay and one for the end of the Gate period (defined by Length, even in FFT. In Level Gate a line is shown only for Delay. You can adjust the position of the green lines by adjusting the gate length and the gate delay. These lines update in the Gate View window as the active function changes, even if the window is not being updated. In Gated LO and Gated Video, these lines are positioned relative to the delay reference line (not relative to 0 time). In Gated FFT, their location is relative to the left edge of the screen.

• A blue line is displayed showing the delay reference, that is, the reference point for the Gate Delay within the Zero Span window. The blue line represents where (in time) the effective location of the gate start would be if the gate were programmed to zero delay.

• The second blue line is labeled "MIN FAST" as shown in the figure above because it represents the minimum Gate Delay for fast Gated LO operation. This line is only displayed in Gated LO. You cannot scroll (knob) or decrement (down key) the Gate Delay to less than that represented by the position of this line, it can only be set below this position manually, although once there it can be moved freely with the knob while below the line.

• A yellow line in the Gated Video case only, is displayed at Blength, where Blength is the display point (bucket) length for the swept trace, which is given by the sweep time for that trace divided by number of Points – 1. So it is referenced to 0 time, not to the delay reference. This line is labeled

NEXT PT (it is not shown in the figure above because the figure above is for Gated LO). The yellow line represents the edge of a display point (bucket). Normally in Gated Video, the bucket length must be selected so that it exceeds the off time of the burst. There is another way to use the analyzer in

Gated Video measurements, and that is to set the bucket width much shorter than the off time of the burst. Then use the Max Hold trace function to fill in "missing" buckets more slowly. This allows you to see some of the patterns of the Gated Video results earlier, though seeing a completely filled-in spectrum later.

Gate View Sweep Time

Controls the sweep time in the Gate View window. To provide an optimal view of the gate signal, the analyzer initializes Gate View Sweep Time based on the current settings of Gate Delay and Gate Length.

Key Path

Sweep/Control, Gate

279

Common Measurement Functions 1

Remote Command

Example

Dependencies

Preset

State Saved

Min

Max

Initial S/W Revision

[:SENSe]:SWEep:EGATe:TIME <time>

[:SENSe]:SWEep:EGATe:TIME?

SWE:EGAT:TIME 500 ms

Gate View Sweep Time is initialized:

On Preset (after initializing delay and length).

Every time the Gate Method is set/changed.

Additionally, in the Swept SA measurement, whenever you do a Preset, or leave Gate View, the analyzer remembers the Gate Delay and Gate Length settings. Then, when returning to Gate View, if the current Gate Delay and/or

Gate Length do not match the remembered values Gate View Sweep Time is re-initialized.

1. Compute the location of the "gate stop" line, which you know is at time t = tmin + GateDelay + GateLength.

519.3 µs

WiMAX OFDMA: 5 ms

GSM/EDGE: 1 ms

Saved in instrument state

1 µs

6000 s

Prior to A.02.00

Gate Delay

Controls the length of time from the time the gate condition goes True until the gate is turned on.

Key Path

Remote Command

Example

Notes

Preset

State Saved

Min

Sweep/Control, Gate

[:SENSe]:SWEep:EGATe:DELay <time>

[:SENSe]:SWEep:EGATe:DELay?

SWE:EGAT:DELay 500ms

SWE:EGAT:DELay?

Units of time are required or no units; otherwise an invalid suffix error will be generated. See error –131.

57.7 us

WiMAX OFDMA: 71 us

GSM/EDGE: 600 us

Saved in instrument state

0.0 us

280

Common Measurement Functions 1

Max

Backwards Compatibility SCPI

Initial S/W Revision

100 s

[:SENSe]:SWEep:TIME:GATE:DELay

Prior to A.02.00

Gate Length

Controls the length of time that the gate is on after it opens.

Key Path

Remote Command

Example

Notes

Dependencies

Preset

State Saved

Min

Max

Backwards Compatibility SCPI

Initial S/W Revision

Sweep/Control, Gate

[:SENSe]:SWEep:EGATe:LENGth <time>

[:SENSe]:SWEep:EGATe:LENGth?

SWE:EGAT:LENG 1

SWE:EGAT:LENG?

Units of time are required or no units; otherwise an invalid suffix will be generated.

Grayed out when Gate Method is set to FFT in which case the label changes to that shown below.

Gate Length

(=1.83/RBW)

2.8 ms vsd 39-1

The key is also grayed out if Gate Control = Level.

461.6 us

WiMAX OFDMA: 50 us

GSM/EDGE: 200 us

Saved in instrument state

100 ns

5 s

[:SENSe]:SWEep:TIME:GATE:LENGth

Prior to A.02.00

Method

This lets you choose one of the three different types of gating.

Not all types of gating are available for all measurements.

Key Path

Remote Command

Sweep/Control, Gate

[:SENSe]:SWEep:EGATe:METHod LO|VIDeo|FFT

[:SENSe]:SWEep:EGATe:METHod?

281

Common Measurement Functions 1

Example

Preset

State Saved

Initial S/W Revision

SWE:EGAT:METH FFT

LO

Saved in instrument state

Prior to A.02.00

LO

When Gate is set to On, the LO sweeps whenever the gate conditions as specified in the Gate menu are satisfied by the signal at the Gate Source.

This form of gating is more sophisticated, and results in faster measurements. With Gated LO, the analyzer only sweeps while the gate conditions are satisfied. This means that a sweep could take place over several gate events. It would start when the gate signal goes true and stop when it goes false, and then continue when it goes true again.

But since the LO is sweeping as long as the gate conditions are satisfied, the sweep typically finishes much more quickly than with Gated Video.

When in zero span, there is no actual sweep performed. But data is only taken while the gate conditions are satisfied. So even though there is no sweep, the gate settings will impact when data is acquired.

Key Path

Dependencies

Sweep/Control, Gate, Method

Key is unavailable when Gate is On and FFT Sweep Type manually selected.

When selected, Sweep Type is forced to Swept and the FFT key in Sweep

Type is grayed out.

LO

Prior to A.02.00

Readback

Initial S/W Revision

Video

When Gate is set to On, the video signal is allowed to pass through whenever the gate conditions as specified in the

Gate menu are satisfied by the signal at the Gate Source.

This form of gating may be thought of as a simple switch, which connects the signal to the input of the spectrum analyzer. When the gate conditions are satisfied, the switch is closed, and when the gate conditions are not satisfied, the switch is open. So we only look at the signal while the gate conditions are satisfied.

With this type of gating, you usually set the analyzer to sweep very slowly. In fact, a general rule is to sweep slowly enough that the gate is guaranteed to be closed at least once per data measurement interval (bucket). Then if the peak detector is used, each bucket will represent the peak signal as it looks with the gate closed.

Key Path

Dependencies

Readback

Initial S/W Revision

Sweep/Control, Gate, Method

Key is unavailable when Gate is On and FFT Sweep Type manually selected.

When selected, Sweep Type is forced to Swept and the FFT key in Sweep

Type is grayed out

Video

Prior to A.02.00

282

Common Measurement Functions 1

FFT

When Gate is set to On, an FFT is performed whenever the gate conditions as specified in the Gate menu are satisfied by the signal at the Gate Source. This is an FFT measurement which begins when the gate conditions are satisfied. Since the time period of an FFT is approximately 1.83/RBW, you get a measurement that starts under predefined conditions and takes place over a predefined period. So, in essence, this is a gated measurement. You have limited control over the gate length but it works in FFT sweeps, which the other two methods do not.

Gated FFT cannot be done in zero span since the instrument is not sweeping. So in zero span the Gated LO method is used. Data is still only taken while the gate conditions are satisfied, so the gate settings do impact when data is acquired.

The Gate Length will be 1.83/RBW.

This is a convenient way to make a triggered FFT measurement under control of an external gating signal.

Key Path

Dependencies

Readback

Initial S/W Revision

Sweep/Control, Gate

Key is unavailable when Gate is On and Swept Sweep Type manually selected.

Key is unavailable when gate Control is set to Level.

When selected, Sweep Type is forced to FFT and the Swept key in Sweep

Type is grayed out

Forces Gate Length to 1.83/RBW

FFT

Prior to A.02.00

Gate Source

The menus under the

Gate Source

key follow the same pattern as those under the

Trigger key

, with the exception that neither

Free Run

nor

Video

are available as Gate Source selections. Any changes to the settings in the setup menus under each Gate Source selection key (for example:

Trigger Level

) also affect the settings under the Trigger menu keys. Note that the selected Trigger Source does not have to match the Gate Source.

Key Path

Remote Command

Dependencies

Preset

Initial S/W Revision

Modified at S/W Revision

Sweep/Control, Gate

[:SENSe]:SWEep:EGATe:SOURce EXTernal1|EXTernal2

|LINE|FRAMe|RFBurst

[:SENSe]:SWEep:EGATe:SOURce?

In some models, there is no second External input. In these models, the

External 2 key is blanked and the EXTernal2 parameter will generate a

“Hardware missing; Not available for this model number” error

EXTernal 1

GSM/EDGE: FRAMe

Prior to A.02.00

A.03.00

283

Common Measurement Functions 1

Control Edge/Level

Sets the method of controlling the gating function from the gating signal.

Edge

In Edge triggering, the gate opens (after the Delay) on the selected edge (for example, positive) of the gate signal and closes on the alternate edge (for example, negative).

Level

In Level triggering, the gate opens (after the Delay) when the gate signal has achieved a certain level and stays open as long as that level is maintained.

Key Path

Remote Command

Example

Dependencies

Preset

State Saved

Backwards Compatibility SCPI

Initial S/W Revision

Sweep/Control, Gate

[:SENSe]:SWEep:EGATe:CONTrol EDGE|LEVel

[:SENSe]:SWEep:EGATe:CONTrol?

SWE:EGAT:CONT EDGE

If the Gate Method is FFT the Control key is grayed out and Edge is selected.

If the Gate Source is TV, Frame or Line, the Control key is grayed out and

Edge is selected.

EDGE

Saved in instrument state

[:SENSe]:SWEep:TIME:GATE:TYPE

Prior to A.02.00

Gate Holdoff

Lets you increase or decrease the wait time after a gate event ends before the analyzer will respond to the next gate signal.

After any Gate event finishes, the analyzer must wait for the sweep system to settle before it can respond to another Gate signal. The analyzer calculates a "wait time," taking into account a number of factors, including RBW and Phase Noise Optimization settings. The goal is to achieve the same accuracy when gated as in ungated operation. The figure below illustrates this concept:

284

Gate Period begins

Gate Period ends

Common Measurement Functions 1

Gate Period begins

Gate

Gate Delay Gate Length

Gate Holdoff

Interval

Gate Delay

Gate

Signal

Gate signal goes true

Gate signal goes true but has no effect as it is during the

Gate Holdoff interval

Gate signal goes true vsd 39-2

When Gate Holdoff is in Auto, the wait time calculated by the analyzer is used. When Gate Time is in

Manual, the user may adjust the wait time, usually decreasing it in order to achieve greater speed, but at the risk of decreasing accuracy.

When the

Method

key is set to

Video

or

FFT

, the

Gate Holdoff

function has no effect.

In measurements that do not support Auto, the value shown when Auto is selected is “---“ and the manually set holdoff is returned to a query.

Key Path

Remote Command

Example

Sweep/Control, Gate

[:SENSe]:SWEep:EGATe:HOLDoff <time>

[:SENSe]:SWEep:EGATe:HOLDoff?

[:SENSe]:SWEep:EGATe:HOLDoff:AUTO OFF|ON|0|1

[:SENSe]:SWEep:EGATe:HOLDoff:AUTO?

SWE:EGAT:HOLD 0.0002

SWE:EGAT:HOLD?

SWE:EGAT:HOLD:AUTO ON

SWE:EGAT:HOLD:AUTO?

285

Common Measurement Functions 1

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

When

Gate Holdoff

is

Auto

, the

Gate Holdoff

key shows the value calculated by the analyzer for the wait time.

Pressing the

Gate Holdoff

key while it is in

Auto

and not selected, causes the key to become selected and allows the user to adjust the value. If the value is adjusted, the setting changes to

Man

.

Pressing the

Gate Holdoff

key, while it is in

Auto

and selected, does not change the value of

Gate Holdoff

, but causes the setting to change to

Man.

Now the user can adjust the value.

Pressing the key while it is in

Man

and selected, cause the value to change back to

Auto

.

Pressing the key while it is in

Man

and not selected, causes the key to become selected and allows the user to adjust the value.

When

Method

is set to

Video

or

FFT

, the

Gate Holdoff

function has no effect.

Auto

Auto/On

Saved in instrument state

1

μsec

1 sec

Prior to A.02.00

Gate Delay Compensation

This function allows you to select an RBW-dependent value by which to adjust the gate delay, to compensate for changes in the delay caused by RBW effects.

Youcan select between uncompensated operation and two types of compensation,

Delay Until RBW

Settled

and

Compensate for RBW Group Delay

.

See

“More Information” on page 287

Key Path

Scope

Remote Command

Example

Sweep/Control, Gate

Meas Global

[:SENSe]:SWEep:EGATe:DELay:COMPensation:TYPE

OFF|SETTled|GDELay

[:SENSe]:SWEep:EGATe:DELay:COMPensation:TYPE?

SWE:EGAT:DEL:COMP:TYPE SETT

SWE:EGAT:DEL:COMP:TYPE?

286

Common Measurement Functions 1

Notes

Preset

State Saved

Range

Readback text

Initial S/W Revision

Although this function is Meas Global, there are some measurements that do not support this function. In those measurements the operation will be

Uncompensated. Going into one of those measurements will not change the

Meas Global selection; it will simply display the grayed-out menu key with

“Uncompensated” showing as the selection. This is a non-forceful grayout, so the SCPI command is still accepted.

If Gate Delay Compensation is not supported at all within a particular mode, the key is not displayed, and if the SCPI command is sent while in a measurement within that mode, an “Undefined Header” error is generated.

Measurements that do not support this function include:

Swept SA

TD-SCDMA mode: Compensate for RBW Group Delay

All other modes: Delay Until RBW Settled

Saved in instrument state

Uncompensated|Delay Until RBW Settled|Compensate for RBW Group

Delay

Uncompensated|Settled|Group Delay

Prior to A.02.00

More Information

Selecting

Uncompensated

means that the actual gate delay is as you sets it.

Selecting

Delay Until RBW Settled

causes the gate delay to be increased above the user setting by an amount equal to 3.06/RBW. This compensated delay causes the GATE START and GATE STOP lines on the display to move by the compensation amount, and the actual hardware gate delay to be increased by the same amount. All the other gate lines (for example, MIN FAST) are unaffected. If the RBW subsequently changes, the compensation is readjusted for the new RBW. The value shown on the

Gate

Delay

key does NOT change.

Delay Until RBW Settled

allows excellent measurements of gated signals, by allowing the IF to settle following any transient that affects the burst. Excellent measurements also require that the analysis region not extend into the region affected by the falling edge of the burst. Thus, excellent measurements can only be made over a width that declines with narrowing RBWs. Therefore, for general purpose compensation, you will still want to change the gate length with changes in RBW even if the gate delay is compensated.

Selecting

Compensate for RBW Group Delay

causes the gate delay to be increased above the user setting by an amount equal to 1.81/RBW. This compensated delay causes the GATE START, GATE

STOP lines on the display to move by the compensation amount, and the actual hardware gate delay to be increased by the same amount. All the other gate lines (for example, MIN FAST) are unaffected. If the RBW subsequently changes, the compensation is readjusted for the new RBW. The value shown on the

Gate Delay

key does NOT change.

Compensate for RBW Group Delay

also includes gate length compensation; the gate length itself is adjusted as necessary to attempt to compensate for delay effects imposed by the RBW.

Compensate for RBW Group Delay

is similar to

Delay Until RBW Settled

, but compensates for the

287

Common Measurement Functions 1

group delay of the RBW filter, rather than the filter settling time. As the RBW gets narrow, this can allow the settling tail of the RBW to affect the beginning part of the gated measurement, and allow the beginning of the RBW settling transient to affect the end of the gated measurement. These two effects are symmetric because the RBW response is symmetric. Because the gate length is not automatically compensated, some users might find this compensation to be more intuitive than compensation for RBW settling.

Min Fast Position Query (Remote Command Only)

This command queries the position of the MIN FAST line, relative to the delay reference (REF) line. See section

“Gate View On/Off” on page 276

. If this query is sent while not in gate view, the MinFast calculation is performed based on the current values of the appropriate parameters and the result is returned. Knowing this value lets you set an optimal gate delay value for the current measurement setup.

Remote Command:

Example:

Initial S/W Revision:

[:SENSe]:SWEep:EGATe:MINFast?

SWE:EGAT:MIN?

Prior to A.02.00

Points

Sets the number of points taken per sweep, and displayed in the traces. The current value of points is displayed parenthetically, next to the sweep time in the lower-right corner of the display. Using more points provides greater resolution; using fewer points compacts the data and decreases the time required to access a trace over the remote interface.

Increasing the number of points does not increase the sweep time; however, it can slightly impact the trace processing time and therefore the overall measurement speed. Decreasing the number of points does not decrease the sweep time, but it may speed up the measurement, depending on the other sweep settings (for example, in FFT sweeps). Fewer points will always speed up the I/O.

Due to minimum sweep rate limitations of the hardware, the minimum sweep time available to the user will increase above its normal value of 1 ms as the number of sweep points increases above 15001.

Changing the number of sweep points has several effects on the analyzer. The sweep time resolution will change. Trace data for all the traces will be cleared and, if Sweep is in Cont, a new trace taken. If any trace is in average or hold, the averaging starts over.

When sweep points is changed, an informational message is displayed, "Sweep points changed, all traces cleared."

Key Path

Remote Command

Example

Dependencies

Sweep

[:SENSe]:SWEep:POINts <integer>

[:SENSe]:SWEep:POINts?

SWE:POIN 5001

SWE:POIN?

Grayed out in measurements that do not support swept

Blanked in modes that do not support swept.

288

Common Measurement Functions 1

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Will be affected by: log sweep

Whenever the number of sweep points change:

All trace data is erased

Any traces with Update Off will also go to Display Off (like going from View to Blank in the older analyzers)

Sweep time is re-quantized

Any limit lines that are on will be updated

If averaging/hold is on, averaging/hold starts over

1001

Saved in instrument state

1

40001

Prior to A.02.00

A.02.00

Abort (Remote Command Only)

This command is used to stop the current measurement. It aborts the current measurement as quickly as possible, resets the sweep and trigger systems, and puts the measurement into an "idle" state. If the analyzer is in the process of aligning when ABORt is sent, the alignment finishes before the abort function is performed. So ABORt does not abort an alignment.

If the analyzer is set for Continuous measurement, it sets up the measurement and initiates a new data measurement sequence with a new data acquisition (sweep) taken once the trigger condition is met.

If the analyzer is set for Single measurement, it remains in the "idle" state until an :INIT:IMM command is received.

Remote Command:

Example:

Notes:

Dependencies:

:ABORt

:ABOR

If :INITiate:CONTinuous is ON, then a new continuous measurement will start immediately; with sweep (data acquisition) occurring once the trigger condition has been met.

If :INITiate:CONTinuous is OFF, then :INITiate:IMMediate is used to start a single measurement; with sweep (data acquisition) occurring once the trigger condition has been met.

For continuous measurement, ABORt is equivalent to the Restart key.

Not all measurements support the abort command.

289

Common Measurement Functions 1

Status Bits/OPC dependencies:

Initial S/W Revision:

The STATus:OPERation register bits 0 through 8 are cleared.

The STATus:QUEStionable register bit 9 (INTegrity sum) is cleared.

Since all the bits that feed into OPC are cleared by the ABORt, the ABORt will cause the *OPC query to return true.

Prior to A.02.00

Trigger

Accesses a menu of keys to control the selection of the trigger source and the setup of each of the trigger sources. The analyzer is designed to allow triggering from a number of different sources, for example,

Free Run, Video, External, RF Burst, and so forth.

The TRIG:SOURCe command (below) will specify the trigger source for the currently selected input

(RF or I/Q). If you change inputs, the new input remembers the trigger source it was last programmed to for the current measurement, and uses that trigger source. You can directly set the trigger source for each input using the TRIGger:RF:SOURce and TRIGger:IQ:SOURce commands (later in this section).

Note the inclusion of the <measurement> parameter in the command below. Because each measurement remembers its own Trigger Source, the command must be qualified with the measurement name. Note that for the Swept SA measurement this is not the case; for backwards compatibility, no <measurement> parameter is used when setting the Trigger Source for the Swept SA measurement.

See

“Trigger Source Presets” on page 291

See

“RF Trigger Source” on page 294

See

“I/Q Trigger Source” on page 296

See

“More Information” on page 297

Remote Command

Example

Initial S/W Revision

Key Path

:TRIGger:<measurement>[:SEQuence]:SOURce

EXTernal1|EXTernal2|IMMediate|LINE|FRAMe|RFBurst|VIDeo|

IF|ALARm|LAN|IQMag|IDEMod|QDEMod|IINPut|QINPut|AIQMag

:TRIGger:<measurement>[:SEQuence]:SOURce?

TRIG:ACP:SOUR EXT1

Selects the external 1 trigger input for the ACP measurement and the selected input

TRIG:SOUR VID

Selects video triggering for the SANalyzer measurement in the Spectrum

Analyzer mode. For SAN, do not use the <measurement> keyword.

Prior to A.02.00

Front-panel key

290

Common Measurement Functions 1

Notes

Dependencies

Preset

Status Bits/OPC dependencies

Backwards Compatibility SCPI

Modified at S/W Revision

Not all measurements have all the trigger sources available to them. Check the trigger source documentation for your specific measurement to see what sources are available.

Not all trigger sources are available for each input. See the

“RF Trigger

Source” on page 294 and

“I/Q Trigger Source” on page 296 commands for

detailed information on which trigger sources are available for each input.

Other trigger-related commands are found in the INITiate and ABORt SCPI command subsystems.

*OPC should be used after requesting data. This will hold off any subsequent changes to the selected trigger source, until after the sweep is completed and the data is returned.

Available ranges and presets can vary from mode to mode.

In some models, there is no second External input. In these models, the

External 2 key is blanked and the EXTernal2 parameter will generate a

“Hardware missing; Not available for this model number” error

See table below

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event occurs and all the applicable trigger criteria have been met). A corresponding pop-up message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec.

This message goes away when a trigger signal appears.

[:SENSe]:<measurement>:TRIGger:SOURce

This backwards compatibility command does not apply to the Swept SA measurement, for that just use :TRIGger:SOURCe

This backwards compatibility command does not apply to the monitor spectrum, log plot and spot frequency measurements at all.

The backwards Compatibility SCPI command,

[:SENSe]:ACPR:TRIGger:SOURce, is provided to support the same functionality as [:SENSe]:ACPr:TRIGger:SOURce (PSA W-CDMA, PSA cdma2000 and PSA 1xEVDO) due to the fact that the ACPr node conflicts with the ACPower node.

In earlier instruments, the parameter IF was used by apps for the video trigger, so using the IF enum selects video triggering.

Sending IF in the command causes VID to be returned to a query.

A.03.00

Trigger Source Presets

Here are the Trigger Source Presets for the various measurements:

Meas Mode Preset for RF Preset for IQ Notes

291

Common Measurement Functions 1

Swept SA SA IMM IQ not supported

CHP

OBW

SA, WCDMA,

C2K, WIMAX

OFDMA,

1xEVDO,

DVB-T/H,

DTMB, LTE,

LTETDD

SA, WCDMA,

C2K, WIMAX

OFDMA,

TD-SCDMA,

1xEVDO, LTE,

LTETDD

IMM

1xEVDO: EXT1 others: IMM

CCDF SA, WCDMA,

C2K,

WIMAXOFDMA

, TD-SCDMA,

1xEV-DO,

DVB-T/H,

DTMB, LTE,

LTETDD

SA, WCDMA,

C2K, LTE:

IMMediate

WIMAXOFDMA

: RFBurst

LTETDD:

BTS: External 1

MS: Periodic

Timer

TD-SCDMA:

BTS: External 1

MS: RFBurst

IMM ACP SA, WCDMA,

C2K, WIMAX

OFDMA,

TD-SCDMA,

1xEVDO,

DVB-T/H,

DTMB, LTE,

LTETDD

Tx Power SA, GSM,

TD-SCDMA

SA, GSM:

RFBurst

TD-SCDMA:

EXTernal

IQ not supported

IQ not supported For 1xEVDO mode, the trigger source is coupled with the gate state, as well as the gate source. When the trigger source changes to RFBurst, External1 or External2, the gate state is set to on, and the gate source is set identically with the trigger source.

When the trigger source changes to

IMMediate, VIDeo, LINE, FRAMe or IF, the gate state is set to off.

TD-SCDMA:

BTS: External 1

MS: IQMag

LTETDD:

BTS: External 1

MS: Periodic

Timer

Others: IMM

For TD-SCDMA:

Trigger source is coupled with radio device.

When radio device changes to BTS, trigger source will be changed to EXTernal1. When radio device changes to MS, trigger source will be set as RFBurst for RF or IQ Mag for

BBIQ.

When TriggerSource is RFBurst or IQ Mag,

Measure Interval is grayed out.

IQ not supported

IMM TD-SCDMA doesn't support the Line and

Periodic Timer parameters.

When the mode is TD-SCDMA, if the Radio

Device is switched to BTS, the value will be changed to External 1 and if the Radio device is switched to MS, the value will be changed to RFBurst

292

Common Measurement Functions 1

SPUR

SEM

CDP

RHO

PCON

QPSK

MON

WAV

PVT

SA, WCDMA,

C2K, WIMAX

OFDMA,

TD-SCDMA,1xE

V-DO, DVB-T/H,

LTE, LTETDD

SA, WCDMA,

C2K, WIMAX

OFDMA,

TD-SCDMA,

1xEVDO,

DVB-T/H,

DTMB, LTE,

LTETDD

IMM

WCDMA

WCDMA

WCDMA

WCDMA, C2K,

1xEVDO

All except SA and

BASIC

SA, WCDMA,

C2K,

TD-SCDMA,

WIMAX

OFDMA, LTE,

LTETDD:

IMMediate

1xEVDO(BTS):

EXTernal1

IMM

IMM

IMM

All except

CDMA1xEVDO:

IMMediate

CDMA1xEVDO:

EXT1

IMM

LTETDD:

BTS: External 1

MS: Periodic

Timer

GSM/EDGE:

RFBurst

All others:

IMMediate

WIMAXOFDMA RFB

IQ not supported

IQ not supported

IMM

IMM

IMM

IMM

IQ not supported

LTETDD:

BTS: External 1

MS: Periodic

Timer

GSM/EDGE:

IQMag

All others:

IMMMediate

IMM

IMM EVM WIMAXOFDMA

, DVB-T/H,

DTMB, LTE,

LTETDD

SPEC BASIC

LOG Plot PN

Spot Freq PN

IMM

IMM

IMM

IMM

IMM

IQ not supported

IQ not supported

LTE, LTETDD supports Free Run, Video and External 1 only.

293

Common Measurement Functions 1

GMSK

PVT

GMSK

PFER

GMSK

ORFS

EDGE

PVT

EDGE

EVM

EDGE

ORFS

Combine d

WCDMA

Combine d

GSM

List

Power

Step

Transmit

On/Off

Power

EDGE/GSM

EDGE/GSM

EDGE/GSM

EDGE/GSM

EDGE/GSM

EDGE/GSM

WCDMA

EDGE/GSM

WCDMA,

EDGE/GSM

LTETDD

RFB

RFB

RFB

RFB

RFB

RFB

IMM

RFB

IMM

IMM

IQMag

IQ not supported

IMM

IQMag

IQ not supported

IQ not supported

IQ not supported

IQ not supported

LTETDD:

BTS: External 1

MS: Periodic

Timer

LTETDD:

BTS: External 1

MS: Periodic

Timer

RF Trigger Source

The

RF Trigger Source

command selects the trigger to be used for the specified measurement when RF is the selected input. The RF trigger source can be queried and changed even while another input is selected, but it is inactive until RF becomes the selected input.

Note the inclusion of the <measurement> parameter in the command below. Because each measurement remembers its own Trigger Source, the command must be qualified with the measurement name. Note that for the Swept SA measurement this is not the case; for backwards compatibility, no <measurement> parameter is used when setting the Trigger Source for the Swept SA measurement.

Remote Command:

:TRIGger:<measurement>[:SEQuence]:RF:SOURce

EXTernal1|EXTernal2|IMMediate|LINE|FRAMe|RFBurst|VIDeo|

IF|ALARm|LAN

:TRIGger:<measurement>[:SEQuence]:RF:SOURce?

294

Common Measurement Functions 1

Example:

Notes:

TRIG:ACP:RF:SOUR EXT1

Selects the external 1 trigger input for the ACP measurement and the RF input

TRIG:RF:SOUR VID

Selects video triggering for the SANalyzer measurement and the RF input.

For SAN, do not use the <measurement> keyword.

Not all measurements have all the trigger sources available to them. Check the trigger source documentation for your specific measurement to see what sources are available.

Not all trigger sources are available for each input. For the

RF Trigger

Source,

the following trigger sources are available:

— IMMediate - free run triggering

— VIDeo - triggers on the video signal level

— LINE - triggers on the power line signal

— EXTernal1 - triggers on an externally connected trigger source marked

“Trigger 1 In” on the rear panel

— EXTernal2 - triggers on an externally connected trigger source marked

“Trigger 2 In” on the front panel. In some models, there is no second

External input. In these models, the External 2 key is blanked and the

EXTernal2 parameter will generate a “Hardware missing; Not available for this model number” error

— RFBurst - triggers on the bursted frame

— FRAMe - triggers on the periodic timer

— IF (video) - same as video, for backwards compatibility only

— ALARm – LXI Alarm

— LAN – LXI LAN event

*OPC should be used after requesting data. This will hold off any subsequent changes to the selected trigger source, until after the sweep is completed and the data is returned.

Available ranges, and presets can vary from mode to mode.

Status Bits/OPC dependencies: The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event occurs and all the applicable trigger criteria have been met). A corresponding pop-up message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec.

This message goes away when a trigger signal appears.

Backwards Compatibility SCPI: In earlier instruments, the parameter IF was used by apps for the video trigger, so using the IF enum selects video triggering.

Sending IF in the query returns the VID enum.

Initial S/W Revision: Prior to A.02.00

295

Common Measurement Functions 1

I/Q Trigger Source

This command selects the trigger to be used for the specified measurement when I/Q (which requires option BBA) is the selected input. The I/Q trigger source can be queried and changed even while another input is selected, but it is inactive until I/Q becomes the selected input.

Remote Command:

Example:

Notes:

Status Bits/OPC dependencies:

Initial S/W Revision:

:TRIGger:<measurement>[:SEQuence]:IQ:SOURce

EXTernal1|EXTernal2|IMMediate|IQMag|IDEMod|QDEMod|IINPu t|QINPut|AIQMag

:TRIGger:<measurement>[:SEQuence]:IQ:SOURce?

TRIG:WAVeform:SOUR IQM

Selects I/Q magnitude triggering for the IQ Waveform measurement and the

I/Q input

Not all measurements have all the trigger sources available to them. Check the trigger source documentation for your specific measurement to see what sources are available.

Not all trigger sources are available for each input. For the

I/Q Trigger

Source,

the following trigger sources are available:

— IMMediate - free run triggering

— EXTernal1 - triggers on an externally connected trigger source on the rear panel

— EXTernal2 - triggers on an externally connected trigger source on the front panel

— IQMag - triggers on the magnitude of the I/Q signal

— IDEMod - triggers on the I/Q signal's demodulated I voltage

— QDEMod - triggers on the I/Q signal's demodulated Q voltage

— IINPut - triggers on the I channel's ADC voltage

— QINPut - triggers on the Q channel's ADC voltage

— AIQMag - triggers on the magnitude of the auxiliary receiver channel I/Q signal

*OPC should be used after requesting data. This will hold off any subsequent changes to the selected trigger source, until after the sweep is completed and the data is returned.

Available ranges, an from mode to mode.d presets can vary

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event occurs and all the applicable trigger criteria have been met). A corresponding pop-up message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec.

This message goes away when a trigger signal appears.

Prior to A.02.00

296

Common Measurement Functions 1

More Information

The trigger menus let you select the trigger source and trigger settings for a sweep or measurement. In triggered operation (basically, any trigger source other than Free Run), the analyzer will begin a sweep or measurement only with the selected trigger conditions are met, generally when your trigger source signal meets the specified trigger level and polarity requirements. (In FFT measurements, the trigger controls when the data acquisition begins for FFT conversion.)

For each of the trigger sources, you may define a set of operational parameters or settings which will be applied when that source is selected as the current trigger source. Examples of these settings are Trigger

Level, Trigger Delay, and Trigger Slope. You may apply different settings for each source; so, for example, you could have a Trigger Level of 1v for External 1 trigger and –10 dBm for Video trigger.

Once you have established the settings for a given trigger source, they generally will remain unchanged for that trigger source as you go from measurement to measurement within a Mode (although the settings do change as you go from Mode to Mode). Furthermore, the trigger settings within a Mode are the same for the Trigger menu, the Gate Source menu, and the Sync Source menu that is part of the Periodic

Timer Trigger Setup menu. That is, if Ext1 trigger level is set to 1v in the Trigger menu, it will appear

as 1v in both the Gate Source and the Sync Source menus. For these reasons the trigger settings commands are not qualified with the measurement name, the way the trigger source commands are.

The settings setup menu can be accessed by pressing the key for the current trigger source a second time.

For example, one press of Video selects the Video trigger as the source. The Video key becomes highlighted and the hollow arrow on the key turns black. Now a second press of the key takes you into the Video Trigger Setup menu.

Trigger Setup Parameters:

The following examples show trigger setup parameters using an external trigger source.

Example 1 illustrates the trigger conditions with negative slope and no trigger occurs during trigger

Holdoff time.

Example 2 illustrates the trigger conditions with positive slope, trigger delay, and auto trigger time.

297

Common Measurement Functions 1

1.5 V

Negative Slope

0.75 V Trigger Level

Trigger Occurs

Example 1:

Trigger Holdoff Time

Positive Slope

(no level selected)

Trigger Delay

Example 2:

Sweep occurs and data processing completes.

Auto Trigger Time

Trigger Occurs

Trigger Delay vsd44

Free Run

Pressing this key, when it is not selected, selects free-run triggering. Free run triggering occurs immediately after the sweep/measurement is initiated.

Key Path

Example

State Saved

Status Bits/OPC dependencies

Initial S/W Revision

Trigger

TRIG:SOUR IMM ! Swept SA measurement

TRIG:<meas>:SOUR IMM ! Measurements other than Swept SA

Saved in instrument state

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event occurs and all the applicable trigger criteria have been met). A corresponding pop-up message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec.

This message goes away when a trigger signal appears.

Prior to A.02.00

Video (IF Envelope)

Pressing this key, when it is not selected, selects the video signal as the trigger. The Video trigger condition is met when the video signal (the filtered and detected version of the input signal, including both RBW and VBW filtering) crosses the video trigger level.

NOTE

When the detector selected for all active traces is the average detector, the video

298

Common Measurement Functions 1

signal for triggering does not include any VBW filtering.

The video trigger level is shown as a labeled line on the display. The line is displayed as long as video is the selected trigger source.

Pressing this key, when it is already selected, accesses the video trigger setup functions.

Key Path

Example

Notes

Dependencies

Trigger

TRIG:SOUR VID ! Swept SA measurement

TRIG:<meas>:SOUR VID ! Measurements other than Swept SA

Log Plot and Spot Frequency measurements do not support Video Trigger

Video trigger is allowed in average detector mode.

State Saved

Status Bits/OPC dependencies

Initial S/W Revision

Saved in instrument state

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event occurs and all the applicable trigger criteria have been met). A corresponding pop-up message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec.

This message goes away when a trigger signal appears.

Prior to A.02.00

Trigger Level

Sets a level for the video signal trigger. When the video signal crosses this level, with the chosen slope, the trigger occurs. This level is displayed with a horizontal line only if Video is the selected trigger source.

Remote Command

Example

Default Unit

Initial S/W Revision

Key Path

:TRIGger[:SEQuence]:VIDeo:LEVel <ampl>

:TRIGger[:SEQuence]:VIDeo:LEVel?

TRIG:VID:LEV –40 dBm depends on the current selected Y axis unit

Prior to A.02.00

Trigger, Video

299

Common Measurement Functions 1

Notes

Couplings

Preset

State Saved

Min

Max

Backwards Compatibility SCPI

When sweep type = FFT, the video trigger uses the amplitude envelope in a bandwidth wider than the FFT width as a trigger source. This might often be useful, but does not have the same relationship between the displayed trace and the trigger level as in swept triggering.

Amplitude Corrections are not taken into account by the Video Trig Level.

For example, if you have given yourself effective gain with an amplitude correction factor, the Video Trigger will not fire until you have dropped the trigger line that far below the displayed signal level, rather than simply dropping it down to the displayed signal level.

Note that other corrections, specifically External Gain and Ref Level Offset, modify the actual trace data as it is taken and therefore ARE taken into account by Trig Level.

This same level is used for the Video trigger source in the Trigger menu and for the Video selection in the Gate Source menu.

Set the Video Trigger Level –25 dBm on Preset. When the Video Trigger

Level becomes the active function, if the value is off screen, set it to either the top or bottom of screen, depending on which direction off screen it was.

Saved in instrument state

–170 dBm

+30 dBm

:TRIGger[:SEQuence]:IF:LEVel

:TRIGger[:SEQuence]:IF:LEVel?

For backward compatibility with VSA/PSA comms apps, we need this alias.

Backwards Compatibility SCPI

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a falling edge.

Key Path

Remote Command

Example

Preset

State Saved

Backwards Compatibility SCPI

Trigger, Video

:TRIGger[:SEQuence]:VIDeo:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:VIDeo:SLOPe?

TRIG:VID:SLOP NEG

POSitive

Saved in instrument state

:TRIGger[:SEQuence]:SLOPe

:TRIGger[:SEQuence]:IF:SLOPe

300

Common Measurement Functions 1

Backwards Compatibility SCPI For backward compatibility, the following commands should update all instances of trigger slope (video/external/line). The query returns the trigger slope setting of the selected trigger source.

:TRIGger[:SEQuence]:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:SLOPe?

For backward compatibility with VSA/PSA comms apps, we need to alias

:TRIGger[:SEQuence]:IF:SLOPe NEGative|POSitive

:TRIGger[:SEQuence]:IF:SLOPe?

Initial S/W Revision Prior to A.02.00

Trig Delay

Controls a time delay during that the analyzer will wait to begin a sweep after meeting the trigger criteria. You can use negative delay to pre-trigger the instrument in the time domain or FFT, but not in swept spans.

Key Path

Remote Command

Example

Notes

Preset

State Saved

Min

Max

Default Unit

Backwards Compatibility SCPI

Trigger, Video

:TRIGger[:SEQuence]:VIDeo:DELay <time>

:TRIGger[:SEQuence]:VIDeo:DELay?

:TRIGger[:SEQuence]:VIDeo:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:VIDeo:DELay:STATe?

TRIG:VID:DEL:STAT ON

TRIG:VID:DEL 100 ms

Video trigger delay may be set to negative values, in time domain, FFT and even swept. It makes intuitive sense in time domain and works well in FFT mode where the bandwidth of the filter before the video trigger is about 1.25 span. In swept spans, negative settings of Trig Delay are treated as a zero setting within the internal hardware and the advisory message "Neg. Trig

Delay unavailable in Swept Mode, zero delay used." is generated when such a delay is set.

Off, 1 us

Saved in instrument state

–150 ms

+500 ms s

:TRIGger[:SEQuence]:IF:DELay

:TRIGger[:SEQuence]:DELay

301

Common Measurement Functions 1

Backwards Compatibility SCPI For backward compatibility with VSA/PSA comms apps, we need to alias

Video trigger to

:TRIGger[:SEQuence]:IF:DELay <time>

:TRIGger[:SEQuence]:IF:DELay?

For backward compatibility, the following commands should update all instances of trigger delay (not including RF Burst). The query returns the video trigger delay settings of the selected trigger source.

:TRIGger[:SEQuence]:DELay <time>

:TRIGger[:SEQuence]:DELay?

:TRIGger[:SEQuence]:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:DELay:STATe?

Also, the legacy ESA command for trigger offset,

TRIGger[:SEQuence]:OFFSet, is supported (see section

“Trigger Offset

(Remote Command Only)” on page 349

). The offset specified by this commands is remembered by the analyzer and added to the video trigger delay whenever the value is sent to the hardware, when in zero span and in a Res

BW >= 1 kHz.

Prior to A.02.00

Initial S/W Revision

Line

Pressing this key, when it is not selected, selects the line signal as the trigger. A new sweep/measurement will start synchronized with the next cycle of the line voltage. Pressing this key, when it is already selected, access the line trigger setup menu.

Key Path

Example

Dependencies

State Saved

Status Bits/OPC dependencies

Trigger

TRIG:SOUR LINE ! Swept SA measurement

TRIG:<meas>:SOUR LINE ! Measurements other than Swept SA

Line trigger is not available when operating from a "dc power source", for example, when the instrument is powered from batteries.

Saved in instrument state

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event occurs and all the applicable trigger criteria have been met). A corresponding pop-up message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec.

This message goes away when a trigger signal appears.

Prior to A.02.00

Initial S/W Revision

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a

302

Common Measurement Functions 1

falling edge.

Key Path

Remote Command

Example

Preset

State Saved

Backwards Compatibility SCPI

Initial S/W Revision

Trigger, Line

:TRIGger[:SEQuence]:LINE:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:LINE:SLOPe?

TRIG:LINE:SLOP NEG

POSitive

Saved in instrument state

:TRIGger[:SEQuence]:SLOPe

(There are SLOPe backward compatibility commands.)

Prior to A.02.00

Trig Delay

Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT, but not in swept spans.

Key Path

Remote Command

Example

Notes

Preset

State Saved

Min

Max

Default Unit

Trigger, Line

:TRIGger[:SEQuence]:LINE:DELay <time>

:TRIGger[:SEQuence]:LINE:DELay?

:TRIGger[:SEQuence]:LINE:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:LINE:DELay:STATe?

TRIG:LINE:DEL:STAT ON

TRIG:LINE:DEL 100 ms

Video trigger delay may be set to negative values, in time domain, FFT and even swept. It makes intuitive sense in time domain and works well in FFT mode where the bandwidth of the filter before the video trigger is about 1.25 span. In swept spans, negative settings of Trig Delay are treated as a zero setting within the internal hardware and the advisory message "Neg. Trig

Delay unavailable in Swept Mode, zero delay used." is generated when such a delay is set.

Off, 1.000 us

Saved in instrument state

–150 ms

500 ms

S

303

Common Measurement Functions 1

Backwards Compatibility SCPI There are DELay backward compatibility commands described in video,

Section

“Trig Delay ” on page 301 )

:TRIGger[:SEQuence]:DELay

(Also, the legacy ESA command for trigger offset,

TRIGger[:SEQuence]:OFFSet, is supported. See section

“Trigger Offset

(Remote Command Only)” on page 349

. The offset specified by this commands is remembered by the analyzer and added to the line trigger delay whenever the value is sent to the hardware, when in zero span and in a Res

BW >= 1 kHz.)

Prior to A.02.00

Initial S/W Revision

External 1

Pressing this key, when it is not selected, selects an external input signal as the trigger. A new sweep/measurement will start when the external trigger condition is met using the external 1 input connector on the rear panel.

Pressing this key, when it is already selected, accesses the external 1 trigger setup menu.

Key Path

Example

State Saved

Status Bits/OPC dependencies

Trigger

TRIG:SOUR EXT1 ! Swept SA measurement

TRIG:<meas>:SOUR EXT1 ! Measurements other than Swept SA

Saved in instrument state

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event occurs and all the applicable trigger criteria have been met). A corresponding pop-up message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec.

This message goes away when a trigger signal appears.

Prior to A.02.00

Initial S/W Revision

Trigger Level

Sets the value where the external 1 trigger input will trigger a new sweep/measurement.

Key Path

Remote Command

Example

Couplings

Preset

Trigger, External 1

:TRIGger[:SEQuence]:EXTernal1:LEVel <level>

:TRIGger[:SEQuence]:EXTernal1:LEVel?

TRIG:EXT1:LEV 0.4 V

This same level is used for the Ext1 trigger source in the Trigger menu, for the

Ext1 selection in the Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for the Ext1 selection in the Gate Source menu.

1.2 V

304

Common Measurement Functions 1

State Saved

Min

Max

Default Unit

Backwards Compatibility SCPI

Initial S/W Revision

Saved in instrument state

–5 V

5 V

V

:TRIGger[:SEQuence]:EXTernal:LEVel

(For backward compatibility, EXTernal should also work.)

Prior to A.02.00

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a falling edge.

Key Path

Remote Command

Example

Couplings

Trigger, External 1

:TRIGger[:SEQuence]:EXTernal1:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:EXTernal1:SLOPe?

TRIG:EXT1:SLOP NEG

This same slope is used in the Ext1 selection for the trigger source in the

Trigger menu and for the period timer sync source (in the Trigger menu and in the Gate Source menu).

POSitive

Saved in instrument state

Preset

State Saved

Backwards Compatibility SCPI For backward compatibility, EXTernal should also work.

Also, there are SLOPe backward compatibility cmds described in Video section

“Trig Slope ” on page 300 :TRIGger[:SEQuence]:SLOPe

Initial S/W Revision Prior to A.02.00

Trig Delay

Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT, but not in swept spans.

Key Path

Remote Command

Trigger, External 1

:TRIGger[:SEQuence]:EXTernal1:DELay <time>

:TRIGger[:SEQuence]:EXTernal1:DELay?

:TRIGger[:SEQuence]:EXTernal1:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:EXTernal1:DELay:STATe?

305

Common Measurement Functions 1

Example

Notes

Preset

State Saved

Min

Max

Default Unit

Backwards Compatibility SCPI

TRIG:EXT1:DEL:STAT ON

TRIG:EXT1:DEL 100 ms

Video trigger delay may be set to negative values, in time domain, FFT and even swept. It makes intuitive sense in time domain and works well in FFT mode where the bandwidth of the filter before the video trigger is about 1.25 span. In swept spans, negative settings of Trig Delay are treated as a zero setting within the internal hardware and the advisory message "Neg. Trig

Delay unavailable in Swept Mode, zero delay used." is generated when such a delay is set.

Off, 1.000 us

Saved in instrument state

–150 ms

+500 ms s

For backward compatibility, EXTernal should also work.

Also, there are DELay backward compatibility commands described in video

section “Trig Delay ” on page 301

:TRIGger[:SEQuence]:DELay

Also, the legacy ESA command for trigger offset,

TRIGger[:SEQuence]:OFFSet, is supported (see section

“Trigger Offset

(Remote Command Only)” on page 349

). The offset specified by this commands is remembered by the analyzer and added to the external1 trigger delay whenever the value is sent to the hardware, when in zero span and in a

Res BW >= 1 kHz.

Prior to A.02.00

Initial S/W Revision

External 2

Pressing this key, when it is not selected, selects an external input signal as the trigger. A new sweep/measurement will start when the external trigger condition is met using the external 2 input connector. The external trigger 2 input connector is on the rear panel.

Pressing this key, when it is already selected, accesses the external 2 trigger setup menu.

Key Path

Example

Dependencies

State Saved

Trigger

TRIG:SOUR EXT2 ! Swept SA measurement

TRIG:<meas>:SOUR EXT2 ! Measurements other than Swept SA

In some models, there is no second External input. In these models, the

External 2 key is blanked and the EXTernal2 parameter will generate a

“Hardware missing; Not available for this model number” error

Saved in instrument state

306

Common Measurement Functions 1

Status Bits/OPC dependencies

Initial S/W Revision

Modified at S/W Revision

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event occurs and all the applicable trigger criteria have been met). A corresponding pop-up message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec.

This message goes away when a trigger signal appears.

Prior to A.02.00

A.03.00

Trigger Level

Sets the value where the external 2 trigger input will trigger a new sweep/measurement.

Key Path

Remote Command

Example

Couplings

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

Trigger, External 2

:TRIGger[:SEQuence]:EXTernal2:LEVel

:TRIGger[:SEQuence]:EXTernal2:LEVel?

TRIG:EXT2:LEV 1.1 V

This same level is used for the Ext2 trigger source in the Trigger menu, for the

Ext2 selection in the Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for the Ext2 selection in the Gate Source menu.

1.2 V

Saved in instrument state

–5 V

5 V

V

Prior to A.02.00

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a falling edge.

Key Path

Remote Command

Example

Couplings

Preset

Trigger, External 2

:TRIGger[:SEQuence]:EXTernal2:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:EXTernal2:SLOPe?

TRIG:EXT2:SLOP NEG

This same slope is used in the Ext2 selection for the trigger source in the

Trigger menu and for the period timer sync source (in the Trigger menu and in the Gate Source menu).

POSitive

307

Common Measurement Functions 1

State Saved

Backwards Compatibility SCPI

Saved in instrument state

Also, there are SLOPe backward compatibility commands described in Video, section :TRIGger[:SEQuence]:SLOPe

Prior to A.02.00

Initial S/W Revision

Trig Delay

Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT, but not in swept spans.

Key Path

Remote Command

Example

Notes

Preset

State Saved

Min

Max

Default Unit

Backwards Compatibility SCPI

Initial S/W Revision

Trigger, External 2

:TRIGger[:SEQuence]:EXTernal2:DELay <time>

:TRIGger[:SEQuence]:EXTernal2:DELay?

:TRIGger[:SEQuence]:EXTernal2:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:EXTernal2:DELay:STATe?

TRIG:EXT2:DEL:STAT ON

TRIG:EXT2:DEL 100 ms

Video trigger delay may be set to negative values, in time domain, FFT and even swept. It makes intuitive sense in time domain and works well in FFT mode where the bandwidth of the filter before the video trigger is about 1.25 span. In swept spans, negative settings of Trig Delay are treated as a zero setting within the internal hardware and the advisory message "Neg. Trig

Delay unavailable in Swept Mode, zero delay used." is generated when such a delay is set.

Off, 1.000 us

Saved in instrument state

–150 ms

500 ms s

Also, there are DELay backward compatibility commands described in video

section “Trig Delay ” on page 301

.

:TRIGger[:SEQuence]:DELay

Also, the legacy ESA command for trigger offset,

TRIGger[:SEQuence]:OFFSet, is supported (see section

“Trigger Offset

(Remote Command Only)” on page 349

). The offset specified by this commands is remembered by the analyzer and added to the external2 trigger delay whenever the value is sent to the hardware, when in zero span and in a

Res BW >= 1 kHz.

Prior to A.02.00

308

Common Measurement Functions 1

RF Burst

Pressing this key, when it is not selected, selects the RF Burst as the trigger. A new sweep/measurement will start when an RF burst envelope signal is identified from the signal at the RF Input connector.

Pressing this key, when it is already selected, accesses the RF Burst trigger setup menu.

In some models, a variety of burst trigger circuitry is available, resulting in various available burst trigger bandwidths. The analyzer automatically chooses the appropriate trigger path based on the hardware configuration and other settings of the analyzer.

Key Path

Example

State Saved

Status Bits/OPC dependencies

Initial S/W Revision

Modified at S/W Revision

Trigger

TRIG:SOUR RFB ! Swept SA measurement

TRIG:<meas>:SOUR RFB ! Measurements other than Swept SA

Saved in instrument state

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event occurs and all the applicable trigger criteria have been met). A corresponding pop-up message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec.

This message goes away when a trigger signal appears.

Prior to A.02.00

A.04.00

Absolute Trigger Level

Sets the absolute trigger level for the RF burst envelope.

Remote Command

Example

Default Unit

Initial S/W Revision

Key Path

Scope

:TRIGger[:SEQuence]:RFBurst:LEVel:ABSolute <ampl>

:TRIGger[:SEQuence]:RFBurst:LEVel:ABSolute?

TRIG:RFB:LEV:ABS 10 dBm sets the trigger level of the RF burst envelope signal to the absolute level of 10 dBm depends on the current selected Y-Axis unit

Prior to A.02.00

Trig, RF Burst

Meas Global

309

Common Measurement Functions 1

Notes

Couplings

Preset

State Saved

Min

Max

Modified at S/W Revision

Sending this command does not switch the setting from relative to absolute; to switch it you need to send the :TRIGger[:SEQuence]:RFBurst:LEVel:TYPE command, below.

Amplitude Corrections are not taken into account by the Absolute Trigger

Level. For example, if you have given yourself effective gain with an amplitude correction factor, the Absolute Trigger will not fire until you have set the trigger level that far below the displayed signal level, rather than simply to the displayed signal level. This is only true for Amplitude

Corrections, not External Gain or Ref Level Offset functions.

This same level is used for the RF Burst trigger source in the Trigger menu, for the RF Burst selection in the Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for the RF Burst selection in the

Gate Source menu

–20 dBm

Saved in state

–200 dBm

100 dBm

A.04.00

Remote Command:

Example:

Preset:

State Saved:

Key Path:

Initial S/W Revision:

Modified at S/W Revision:

:TRIGger[:SEQuence]:RFBurst:LEVel:TYPE

ABSolute|RELative

:TRIGger[:SEQuence]:RFBurst:LEVel:TYPE?

TRIG:RFB:LEV:TYPE REL sets the trigger level type of the RF burst trigger to Relative.

ABSolute

Saved in State

Trig, RF Burst

Prior to A.02.00

A.04.00

Relative Trigger Level

Sets the relative trigger level for the RF burst envelope.

In some models, the relative burst trigger function is implemented in hardware. In other models, without the advanced triggering hardware required, the relative burst trigger function is implemented in software in some measurements, and is unavailable in other measurements.

When implemented in software, the relative RF Burst trigger function is implemented as follows:

The measurement starts with the absolute RF Burst trigger setting. If it can not get a trigger with that level, auto trigger fires and the acquisition starts anyway. After the acquisition, the measurement searches for the peak in the acquired waveform and saves it.

310

Common Measurement Functions 1

Now, in the next cycle of the measurement, the measurement determines a new absolute RF Burst level based on the peak value from the first measurement and the Relative RF Burst Trigger Level (always 0 or negative dB) set by the user. The following formula is used: absolute RF Burst level = peak level of the previous acquisition + relative RF Burst level

If the new absolute RF Burst level differs from the previous by more than 0.5 dB, the new level is sent to the hardware; otherwise it is not updated (to avoid slowing down the acquisition)

Steps 2 and 3 repeat for subsequent measurements.

Remote Command

Example

Default Unit

Initial S/W Revision

Key Path

Scope

Notes

Dependencies

Preset

State Saved

Min

Max

Backwards Compatibility SCPI

Modified at S/W Revision

:TRIGger[:SEQuence]:RFBurst:LEVel:RELative <rel_ampl>

:TRIGger[:SEQuence]:RFBurst:LEVel:RELative?

TRIG:RFB:LEV:REL –10 dB sets the trigger level of the RF burst envelope signal to the relative level of

–10 dB dB or dBc

Prior to A.02.00

Trig, RF Burst

Meas Global

Sending this command does not switch the setting from absolute to relative; to switch it you need to send the :TRIGger[:SEQuence]:RFBurst:LEVel:TYPE command, above.

The relative trigger level is not available in some measurements. In those measurements the RELative parameter, and the

:TRIGger[:SEQuence]:RFBurst:LEVel:TYPE command (above), should generate an error if sent.

This key is grayed out and Absolute Trigger Level selected if the required hardware is not present in your analyzer and the current measurement does not support Relative triggering.

–6 dB

GSM: –25 dB

Saved in state

–45 dB

0 dB

:TRIGger[:SEQuence]:RFBurst:LEVel

Is aliased to:

:TRIGger[:SEQuence]:RFBurst:LEVel:RELative because the PSA had ONLY relative burst triggering

A.04.00

311

Common Measurement Functions 1

Trigger Slope

It is set positive to trigger on a rising edge and negative to trigger on a falling edge.

Key Path

Remote Command

Example

Couplings

Preset

State Saved

Backwards Compatibility SCPI

Initial S/W Revision

Trigger, RF Burst

:TRIGger[:SEQuence]:RFBurst:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:RFBurst:SLOPe?

TRIG:RFB:SLOP NEG

This same slope is used in the RF Burst selection for the trigger source in the

Trigger menu and for the period timer sync source (in the Trigger menu and in the Gate Source menu).

POSitive

Saved in instrument state

Also, there are SLOPe backward compatibility commands described in Video

section “Trig Slope ” on page 300

:TRIGger[:SEQuence]:SLOPe

Prior to A.02.00

Trig Delay

Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT, but not in swept spans.

Key Path

Remote Command

Example

Notes

Preset

State Saved

Min

Trigger, RF Burst

:TRIGger[:SEQuence]:RFBurst:DELay <time>

:TRIGger[:SEQuence]:RFBurst:DELay?

:TRIGger[:SEQuence]:RFBurst:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:RFBurst:DELay:STATe?

TRIG:RFB:DEL:STAT ON

TRIG:RFB:DEL 100 ms

Video trigger delay may be set to negative values, in time domain, FFT and even swept. It makes intuitive sense in time domain and works well in FFT mode where the bandwidth of the filter before the video trigger is about 1.25 span. In swept spans, negative settings of Trig Delay are treated as a zero setting within the internal hardware and the advisory message "Neg. Trig

Delay unavailable in Swept Mode, zero delay used." is generated when such a delay is set.

Off, 1.000 us

Saved in instrument state

–150 ms

312

Common Measurement Functions 1

Max

Default Unit

500 ms s

Backwards Compatibility SCPI Also, there are DELay backward compatibility commands described in video section

“Trig Delay ” on page 301 .

:TRIGger[:SEQuence]:DELay

Initial S/W Revision Prior to A.02.00

Periodic Timer (Frame Trigger)

Pressing this key, when it is not selected, selects the internal periodic timer signal as the trigger.

Triggering occurrences are set by the Period parameter, which is modified by the Sync Source and

Offset. Pressing this key, when it is already selected, accesses the periodic timer trigger setup functions.

If you do not have a sync source selected (it is Off), then the internal timer will not be synchronized with any external timing events.

Key Path

Example

State Saved

Readback

Status Bits/OPC dependencies

Initial S/W Revision

Trigger

TRIG:SOUR FRAM ! Swept SA measurement

TRIG:<meas>:SOUR FRAM ! Measurements other than Swept SA

Saved in instrument state

[Sync: <value of Sync Source>], for example, [Sync: External 1]

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event occurs and all the applicable trigger criteria have been met). A corresponding pop-up message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec.

This message goes away when a trigger signal appears.

Prior to A.02.00

Periodic Timer Triggering:

This feature selects the internal periodic timer signal as the trigger. Trigger occurrences are set by the

Periodic Timer parameter, which is modified by the Sync Source and Offset.

The figure below shows the action of the periodic timer trigger. Before reviewing the figure, we’ll explain some uses for the periodic trigger.

A common application is measuring periodic burst RF signals for which a trigger signal is not easily available. For example, we might be measuring a TDMA radio which bursts every 20 ms. Let’s assume that the 20 ms period is very consistent. Let’s also assume that we do not have an external trigger source available that is synchronized with the period, and that the signal-to-noise ratio of the signal is not high enough to provide a clean RF burst trigger at all of the analysis frequencies. For example, we might want to measure spurious transmissions at an offset from the carrier that is larger than the bandwidth of the RF burst trigger. In this application, we can set the Periodic Timer to a 20.00 ms period and adjust the offset from that timer to position our trigger just where we want it. If we find that the 20.00 ms is not exactly right, we can adjust the period slightly to minimize the drift between the period timer and the signal to be

313

Common Measurement Functions 1

measured.

A second way to use this feature would be to use Sync Source temporarily, instead of Offset. In this case, we might tune to the signal in a narrow span and use the RF Burst trigger to synchronize the periodic timer. Then we would turn the sync source off so that it would not mis-trigger. Mis-triggering can occur when we are tuned so far away from the RF burst trigger that it is no longer reliable.

A third example would be to synchronize to a signal that has a reference time element of much longer period than the period of interest. In some CDMA applications, it is useful to look at signals with a short periodicity, by synchronizing that periodicity to the "even-second clock" edge that happens every two seconds. Thus, we could connect the even-second clock trigger to Ext1 and use then Ext1 as the sync source for the periodic timer.

The figure below illustrates this third example. The top trace represents the even-second clock. It causes the periodic timer to synchronize with the leading edge shown. The analyzer trigger occurs at a time delayed by the accumulated offset from the period trigger event. The periodic timer continues to run, and triggers continue to occur, with a periodicity determined by the analyzer time base. The timer output

(labeled "late event") will drift away from its ideal time due to imperfect matching between the time base of the signal being measured and the time base of the analyzer, and also because of imperfect setting of the period parameter. But the synchronization is restored on the next even-second clock event.

("Accumulated offset" is described in the in the Offset function section.)

Ext 1

Even-second Clock

Late Event

Period

Timer

Period

Trigger

Accumulated Offset vsd45

Period

Sets the period of the internal periodic timer clock. For digital communications signals, this is usually set to the frame period of your current input signal. In the case that sync source is not set to OFF, and the external sync source rate is changed for some reason, the periodic timer is synchronized at the every external synchronization pulse by resetting the internal state of the timer circuit.

Key Path

Remote Command

Example

Trigger, Periodic Timer

:TRIGger[:SEQuence]:FRAMe:PERiod <time>

:TRIGger[:SEQuence]:FRAMe:PERiod?

TRIG:FRAM:PER 100 ms

314

Common Measurement Functions 1

Notes

Dependencies

Couplings

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

Internal hardware resolution is 1/ (30 MHz 216) or about 0.5 ps. The manual

UI resolution should be 1 ps. The remote command can be sent with more resolution but must be quantized to the hardware resolution when applied.

Therefore, Joe Gorin proposed an internal software resolution smaller than 0.5 ps. Let’s choose 0.1 ps.

The invalid data indicator turns on when the period is changed, until the next sweep/measurement completes.

The same period is used in the Gate Source selection of the period timer.

20 ms

GSM: 4.615383

Saved in instrument state

100.000 ns

559.0000 ms

S

Prior to A.02.00

Offset

Adjusts the accumulated offset between the periodic timer events and the trigger event. Adjusting the accumulated offset is different than setting an offset, and requires explanation.

The periodic timer is usually not synchronized with any external events, so the timing of its output events has no absolute meaning. Since the timing relative to external events (RF signals) is important, you need to be able to adjust (offset) it. However, you have no direct way to see when the periodic timer events occur. All that you can see is the trigger timing. When you want to adjust the trigger timing, you will be changing the internal offset between the periodic timer events and the trigger event. Because the absolute value of that internal offset is unknown, we will just call that the accumulated offset. Whenever the Offset parameter is changed, you are changing that accumulated offset. You can reset the displayed offset using Reset Offset Display. Changing the display does not change the value of the accumulated offset, and you can still make additional changes to accumulated offset.

To avoid ambiguity, we define that an increase in the "offset" parameter, either from the knob or the

SCPI adjust command, serves to delay the timing of the trigger event.

Key Path

Remote Command

Example

Trigger, Periodic Timer

:TRIGger[:SEQuence]:FRAMe:OFFSet <time>

:TRIGger[:SEQuence]:FRAMe:OFFSet?

TRIG:FRAM:OFFS 1.2 ms

315

Common Measurement Functions 1

Notes

Notes

Dependencies

Couplings

The front panel interface (for example, the knob), and this command, adjust the accumulated offset, which is shown on the active function display.

However, the actual amount sent to the hardware each time the offset is updated is the delta value, that is, the current accumulated offset value minus the previous accumulated offset value. Note that the accumulated offset value is essentially arbitrary; it represents the accumulated offset from the last time the offset was zeroed (with the Reset Offset Display key).

Note that this command does not change the period of the trigger waveform.

Note also that Offset is used only when the sync source is set to OFF, otherwise delay is used, see section

“Trig Delay” on page 320

.

An increase in the "offset" parameter, either from the knob or the SCPI adjust command, serves to delay the timing of the trigger event.

When the SCPI command is sent the value shown on the key (and the Active

Function, if this happens to be the active function) is updated with the new value. However, the actual amount sent to the hardware is the delta value, that is, the current accumulated offset value minus the previous accumulated offset value.

The SCPI query simply returns the value currently showing on the key.

The invalid data indicator turns on when the offset is changed, until the next sweep/measurement completes.

The same offset is used in the Gate Source selection of the period timer.

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

0 s

Saved in instrument state

–10.000 s

10.000 s

S

Prior to A.02.00

Offset Adjust (Remote Command Only)

This remote command does not work at all like the related front panel keys. This command lets you advance the phase of the frame trigger by the amount you specify.

It does not change the period of the trigger waveform. If the command is sent multiple times, it advances the phase of the frame trigger an additional amount each time it is sent. Negative numbers are permitted.

Remote Command:

Example:

Notes:

:TRIGger[:SEQuence]:FRAMe:ADJust <time>

TRIG:FRAM:ADJ 1.2 ms

Note also that Offset is used only when the sync source is set to OFF, otherwise delay is used, see section

“Trig Delay” on page 320

An increase in the "offset" parameter, either from the knob or the SCPI adjust command, serves to delay the timing of the trigger event.

316

Common Measurement Functions 1

Notes:

Dependencies:

Couplings:

Preset:

State Saved:

Min:

Max:

Default Unit:

Initial S/W Revision:

The front panel interface (for example, the knob) and the :TRIG:FRAM:OFFS command adjust the accumulated offset, which is shown on the active function display. However, the actual amount sent to the hardware is the delta value, that is, the current offset value minus the previous offset value.

When the SCPI command is sent the value shown on the key (and the Active

Function, if this happens to be the active function) is updated by increasing it

(or decreasing it if the value sent is negative) by the amount specified in the

SCPI command.

This is a "command only" SCPI command, with no query.

The invalid data indicator turns on when the offset is changed, until the next sweep/measurement completes.

The same offset is used in the Gate Source selection of the period timer.

0 s

Saved in instrument state

–10.000 s

10.000 s

S

Prior to A.02.00

Reset Offset Display

Resets the value of the periodic trigger offset display setting to 0.0 seconds. The current displayed trigger location may include an offset value defined with the

Offset

key. Pressing this key redefines the currently displayed trigger location as the new trigger point that is 0.0 s offset. The

Offset

key can then be used to add offset relative to this new timing.

Key Path

Remote Command

Example

Initial S/W Revision

Trigger, Periodic Timer

:TRIGger[:SEQuence]:FRAMe:OFFSet:DISPlay:RESet

TRIG:FRAM:OFFS:DISP:RES

Prior to A.02.00

Sync Source

Selects a signal source for you to synchronize your periodic timer trigger to, otherwise you are triggering at some arbitrary location in the frame. Synchronization reduces the precision requirements on the setting of the period.

For convenience you may adjust the level and slope of the selected sync source in a conditional branch setup menu accessed from the Sync Source menu. Note that these settings match those in the Trigger and Gate Source menus; that is, each trigger source has only one value of level and slope, regardless of which menu it is accessed from.

Key Path

Trigger, Periodic Timer

317

Common Measurement Functions 1

Remote Command

Example

Dependencies

Preset

State Saved

Readback

Initial S/W Revision

Modified at S/W Revision

:TRIGger[:SEQuence]:FRAMe:SYNC

EXTernal1|EXTernal2|RFBurst|OFF

:TRIGger[:SEQuence]:FRAMe:SYNC?

TRIG:FRAM:SYNC EXT2

In some models, there is no second External input. In these models, the

External 2 key is blanked and the EXTernal2 parameter will generate a

“Hardware missing; Not available for this model number” error

Off

GSM/EDGE: RFBurst

Saved in instrument state

The current setting is read back to this key and it is also Readback to the previous

Periodic Timer

trigger key.

Prior to A.02.00

A.03.00

Off

Turns off the sync source for your periodic trigger. With the sync source off, the timing will drift unless the signal source frequency is locked to the analyzer frequency reference.

Key Path

Example

Readback

Initial S/W Revision

Trigger, Periodic Timer, Sync Source

TRIG:FRAM:SYNC OFF

Off

Prior to A.02.00

External 1

Pressing this key, when it is not selected, selects the external input port that you will use for the periodic trigger synchronization. Pressing this key, when it is already selected, accesses the external 1 sync source setup menu.

Key Path

Example

Couplings

Readback

Initial S/W Revision

Trigger, Periodic Timer, Sync Source

TRIG:FRAM:SYNC EXT

Same as External 1 trigger source.

External 1

Prior to A.02.00

Trigger Level

Sets the value where the signal at the external 1 trigger input will synchronize with the periodic timer trigger. This

same level is used in the Ext1 trigger source in the Trigger menu. See section “Trigger Level ” on page 304 for

information on this key and the SCPI command.

318

Common Measurement Functions 1

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a falling edge.

This same value is used in the Ext1 trigger source in the Trigger menu. See section

“Trig Slope ” on page 305 for

information on this key and the SCPI command

External 2

Pressing this key, when it is not selected, selects the external input port that you will use for the periodic frame trigger synchronization.

Pressing this key, when it is already selected, accesses the external 2 sync source setup menu.

Key Path

Example

Dependencies

Couplings

Readback

Initial S/W Revision

Modified at S/W Revision

Trigger, Periodic Timer, Sync Source

TRIG:FRAM:SYNC EXT2

In some models, there is no second External input. In these models, the

External 2 key is blanked and the EXTernal2 parameter will generate a

“Hardware missing; Not available for this model number” error

Same as External 2 trigger source.

External 2

Prior to A.02.00

A.03.00

Trigger Level

Sets the value where the signal at the external 2 trigger input will synchronize with the periodic timer trigger. This

same level is used in the Ext2 trigger source in the Trigger menu. See section “Trigger Level ” on page 307

for information on this key and the SCPI command.

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a falling edge.

This same value is used in the Ext2 trigger source in the Trigger menu. See section

“Trig Slope ” on page 307 for

information on this key and the SCPI command

RF Burst

Pressing the key once selects the RF burst envelope signal to be used for the periodic timer trigger synchronization.

Press the key a second time to access the RF burst sync source setup menu.

Key Path

Example

Couplings

Readback

Initial S/W Revision

Trigger, Periodic Timer, Sync Source

TRIG:FRAM:SYNC RFB

Same as RF Burst trigger source.

RF Burst

Prior to A.02.00

Trigger Level

Sets the trigger level to be used for the RF Burst trigger. This same level is used in the RF Burst trigger source in

319

Common Measurement Functions 1

the Trigger menu. See section “Absolute Trigger Level” on page 309 for information on this key and the SCPI

command.

Trig Slope

Controls the RF Burst trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a falling edge. This same value is used in the RF Burst trigger source in the Trigger menu. See section

“Trigger

Slope ” on page 312 for information on this key and the SCPI command

Trig Delay

This setting delays the measurement timing relative to the Periodic Timer.

Key Path

Remote Command

Notes

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

Trigger, Periodic Timer

:TRIGger[:SEQuence]:FRAMe:DELay <time>

:TRIGger[:SEQuence]:FRAMe:DELay?

:TRIGger[:SEQuence]:FRAMe:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:FRAMe:DELay:STATe?

Note that delay is used when the sync source is not set to OFF. If the sync source is set to OFF, offset is used, see section 5.3.7.2.

Off, 1.000 us

Saved in instrument state

–150 ms

+500 ms s

Prior to A.02.00

Sync Holdoff

Sync Holdoff specifies the duration that the sync source signal must be kept false before the transition to true to be recognized as the sync timing. The periodic timer phase is aligned when the sync source signal becomes true, after the Holdoff time is satisfied.

A holdoff of 2 ms will work with most WiMAX signals, but there may be cases where the burst off duration is less than 1 ms and this value will need to be changed.

Key Path

Remote Command

Preset

State Saved

Trigger, Periodic Timer

:TRIGger[:SEQuence]:FRAMe:SYNC:HOLDoff <time>

:TRIGger[:SEQuence]:FRAMe:SYNC:HOLDoff?

:TRIGger[:SEQuence]:FRAMe:SYNC:HOLDoff:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:FRAMe:SYNC:HOLDoff:STATe?

On, 1.000 ms

Saved in instrument state

320

Common Measurement Functions 1

Min

Max

Default Unit

Initial S/W Revision

0 ms

+500 ms s

Prior to A.02.00

LXI Trigger

Pressing this key when it is not selected selects the LXI system as the trigger. Pressing the key when it is already selected accesses the LXI trigger type selection menu, where either LAN Event or Alarm can be chosen. The key is annotated to display which of the two is currently selected.

NOTE

For information about setting up measurements using LXI, refer to the

Programmer’s Guide located in your analyzer at: C:/Program Files/Agilent/Signal

Analysis/Help/Bookfiles/x_series_prog.pdf. It is also available by selecting the

"Additional Documentation" page of the Help.

Key Path

Preset

State Saved

Readback

Trigger

ON

Saved in instrument state

The LXI trigger source that becomes active when this key is selected is displayed. The possible values are "LAN Event" and "Alarm"

Prior to A.02.00

Initial S/W Revision

LAN Event

Pressing this key when it is not selected selects the LAN event system as the LXI trigger. A new sweep/measurement starts when the pre-configured LAN message arrives if the LXI trigger is selected

(see “LXI Trigger” on page 321 ). Pressing this key when it is already selected accesses the LAN trigger

setup menu.

NOTE

Pressing this key causes Enabled LXI Alarm Triggers to be ignored, since the

Trigger source is changed to LXI LAN Event.

Key Path

Example

Preset

State Saved

Trigger, LXI Trigger

TRIG:SOUR LAN ! Swept SA measurement

TRIG:<meas>:SOUR LAN ! Measurements other than Swept SA

ON

Saved in instrument state

321

Common Measurement Functions 1

Status Bits/OPC dependencies The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event occurs and all the applicable trigger criteria have been met). A corresponding pop-up message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec.

This message goes away when a trigger signal appears.

Prior to A.02.00

Initial S/W Revision

Disable All

Sets the Enable parameter of every member of the LXI LAN Event list to OFF.

Key Path

Remote Command

Example

Initial S/W Revision

Trigger, LXI Trigger, LAN Event

:TRIGger[:SEQuence]:LXI:LAN:DISable:ALL

:TRIG:LXI:LAN:DIS:ALL

Prior to A.02.00

LAN Event List

After selecting LAN as the trigger source, the user is presented with a list of LXI Trigger LAN Events to be configured. By default, LAN0-LAN7 are available. Using the TRIG:LXI:LAN:ADD and TRIG:LXI:LAN:REM commands, the size of this list can be changed arbitrarily. Pressing a LAN event branches to that event’s setup menu.

Key Path

Remote Command

Example

Preset

State Saved

Readback

Trigger, LXI Trigger, LAN Event

:TRIGger[:SEQuence]:LXI:LAN:LIST?

:TRIG:LXI:LAN:LIST? !Returns the complete list of Trigger LAN Events which is, at minimum: "LAN0", "LAN1", "LAN2", "LAN3", "LAN4",

"LAN5", "LAN6", "LAN7"

"LAN0", "LAN1", "LAN2", "LAN3", "LAN4", "LAN5", "LAN6", "LAN7"

Saved in instrument state

Displays the value of the LXI Trigger LAN Event parameter

(Enabled|Disabled).

Prior to A.02.00

Initial S/W Revision

Detection

Pressing this key accesses the Trigger Detection menu.

Selecting "Rise" causes the instrument to trigger on the receipt of a signal low LAN Event followed by a signal high LAN Event.

Selecting "Fall" caused the instrument to trigger on the receipt of a signal high LAN Event followed by a signal low

LAN Event.

Selecting "High" causes the instrument to trigger on every signal high LAN Event.

322

Common Measurement Functions 1

Selecting "Low" causes the instrument to trigger on every signal low LAN Event.

Key Path

Remote Command

Example

Notes

Preset

State Saved

Range

Readback

Initial S/W Revision

Trigger, LXI Trigger, LAN Event, <lanEvent>

:TRIGger[:SEQuence]:LXI:LAN[:SET]:DETection "LANEVENT",

HIGH|LOW|RISE|FALL

:TRIG:LXI:LAN:DET "LAN0",HIGH

If a non existent LAN event is passed in the lanEvent argument, the command is ignored

HIGH

Saved in instrument state

HIGH | LOW | RISE | FALL

Currently selected detection type

Prior to A.02.00

Key Path

Remote Command

Example

Notes

Preset

State Saved

Range

Readback

Initial S/W Revision

Trigger, LXI Trigger, LAN Event, <lanEvent>

:TRIGger[:SEQuence]:LXI:LAN[:SET]:DETection? "LANEVENT"

:TRIG:LXI:LAN:DET? "LAN0"?

If a non existent LAN event is passed in the lanEvent argument, the command is ignored

HIGH

Saved in instrument state

HIGH | LOW | RISE | FALL

Currently selected detection type

Prior to A.02.00

Delay

Sets the amount of delay that should pass between receiving a LXI Trigger LAN Event Trigger and the trigger action. A Delay of 0.0 s indicates that the instrument will trigger as soon as possible after receiving the proper LXI

LAN Event.

Key Path

Remote Command

Example

Preset

State Saved

Range

Trigger, LXI Trigger, LAN Event, <lanEvent>

:TRIGger[:SEQuence]:LXI:LAN[:SET]:DELay

"LANEVENT",<time>

:TRIG:LXI:LAN:DEL "LAN0",5S

0.0 s

Saved in instrument state

0.0 – 1.7976931348623157 x 10308 (Max Double)

323

Common Measurement Functions 1

Initial S/W Revision Prior to A.02.00

Key Path

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Trigger, LXI Trigger, LAN Event, <lanEvent>

:TRIGger[:SEQuence]:LXI:LAN[:SET]:DELay? "lanEvent"

:TRIG:LXI:LAN:DEL? "LAN0"

0.0 s

Saved in instrument state

0.0 – 1.7976931348623157 x 10308 (Max Double)

Prior to A.02.00

Enabled/Disabled

When the Trigger Source is set to LXI Trigger LAN Event, the instrument triggers upon receiving any event from the LXI Trigger LAN Event List whose Enabled parameter is set to ON.

If the Enabled parameter is set to OFF, the event is ignored.

Key Path

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Trigger, LXI Trigger, LAN Event, <lanEvent>

:TRIGger[:SEQuence]:LXI:LAN[:SET]:ENABled

"LANEVENT",ON|OFF|1|0

:TRIG:LXI:LAN:ENAB "LAN0",ON

OFF

Saved in instrument state

OFF|ON|0|1

Prior to A.02.00

Key Path

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Trigger, LXI Trigger, LAN Event, <lanEvent>

:TRIGger[:SEQuence]:LXI:LAN[:SET]:ENABled? "LANEVENT"

:TRIG:LXI:LAN:ENAB? "LAN0"

OFF

Saved in instrument state

OFF|ON|0|1

Prior to A.02.00

Add (Remote Command Only)

Adds the provided string to the list of possible LAN events to trigger on. As new LAN events are added, keys are generated in the LAN source menu. New key panels are generated as the number of possible LAN events increases past a multiple of six, and the "More" keys are updated to reflect the new number of key panels in the LAN source

324

Common Measurement Functions 1

menu.

Remote Command:

Example:

Notes:

State Saved:

Range:

Initial S/W Revision:

:TRIGger[:SEQuence]:LXI:LAN:ADD "LANEVENT"

:TRIG:LXI:LAN:ADD "LANEVENT"

The maximum length of the string is 16 characters.

Longer strings are concatenated to 16 characters and added.

No event is added if the LAN Event already exists.

This command modifies the LXI Trigger LAN Event List Parameter.

Saved in instrument state

Uppercase, Lowercase, Numeric, Symbol except for comma or semicolon

Prior to A.02.00

Remove (Remote Command Only)

Removes the provided string from the list of possible LAN events to trigger on. As LAN events are removed, keys are removed from the LAN source menu. Key panels are removed as the number of possible LAN events decreases past a multiple of six, and the "More" keys are updated to reflect the new number of key panels in the LAN source menu. It is not possible to remove the "LAN0" – "LAN7" events.

Remote Command:

Example:

Notes:

State Saved:

Range:

Initial S/W Revision:

:TRIGger[:SEQuence]:LXI:LAN:REMove[:EVENt] "LANEVENT"

:TRIG:LXI:LAN:REM "LANEVENT"

The maximum length of the string is 16 characters.

Longer strings are concatenated and the corresponding LAN Event is removed.

Nothing happens if the LAN event does not exist.

This command modifies the LXI Trigger LAN Event List Parameter.

No

Uppercase, Lowercase, Numeric, Symbol except for comma or semicolon

Prior to A.02.00

Remove All (Remote Command Only)

Clears the list of customer added LAN events that can cause the instrument to trigger. Events LAN0-LAN7 are not affected. As LAN events are removed, keys are removed from the LAN source menu. Key panels are removed as the number of possible LAN events decreases past a multiple of six, and the "More" keys are updated to reflect the new number of key panels in the LAN source menu.

It is not possible to remove the "LAN0" – "LAN7" events.

Remote Command:

Example:

:TRIGger[:SEQuence]:LXI:LAN:REMove:ALL

:TRIG:LXI:LAN:REM:ALL

325

Common Measurement Functions 1

Notes:

Initial S/W Revision:

This command modifies the LXI Trigger LAN Event List Parameter.

Prior to A.02.00

Event Filter (Remote Command Only)

Only LXI Trigger LAN Events coming from hosts matching the filter string are processed. There is no front panel access to this command

The syntax for specifying a filter is as follows:

Filter == ( [host[:port]] | [ALL[:port]] ) [,Filter]

Specifying an empty string means that LXI trigger packets are accepted as a Trigger from any port on any host on the network via either TCP or UDP.

Specifying only the port means that any host communicating over that port can send events.

Specifying ALL indicates that UDP multicast packets are accepted if they are directed to the Internet Assigned

Numbers Authority (IANA) assigned multicast address on the IANA assigned default port, or the designated port if specified.

Examples:

"192.168.0.1:23"

"agilent.com, soco.agilent.com"

"agilent.com:80, 192.168.0.1"

• The TRIGger:LXI:LAN:FILTer command applies only to trigger events and will have no effect on state events, even when both are tied to the same event name (like "LAN0"). Similarly, the LXI:EVENt:INPut:LAN:FILTer command applies only to state events and will have no effect on trigger events.

Remote Command:

Example:

Notes:

Preset:

State Saved:

Range:

Initial S/W Revision:

:TRIGger[:SEQuence]:LXI:LAN[:SET]:FILTer

"LANEVENT","filterString"

:TRIGger[:SEQuence]:LXI:LAN[:SET]:FILTer?

:TRIG:LXI:LAN:FILT "LAN0","agilent.com"

:TRIG:LXI:LAN:FILT?

The maximum length of the string is 45 characters. Nothing happens if the

LAN event does not exist.

“” (empty string)

Saved in instrument state

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

Count (Remote Command Only)

Returns the number of items in the LXI Trigger LAN Event List.

Remote Command:

:TRIGger[:SEQuence]:LXI:LAN:COUNt?

326

Common Measurement Functions 1

Example:

Initial S/W Revision:

:TRIG:LXI:LAN:COUN?

Prior to A.02.00

Identifier (Remote Command Only)

Sets the string that is expected to arrive over the LAN for a given Trigger LAN Event to occur. The Identifier is variable to allow for easier system debugging.

Remote Command:

Example:

Notes:

State Saved:

Range:

Initial S/W Revision:

:TRIGger[:SEQuence]:LXI:LAN[:SET]:IDENtifier

"LANEVENT","identifier"

:TRIGger[:SEQuence]:LXI:LAN[:SET]:IDENtifier?

"LANEVENT"

:TRIG:LXI:LAN:IDEN "LAN0","debugstring"

The maximum length of the string is 16 characters.

Nothing happens if the LAN event does not exist.

The default value is that the identifier is equivalent to the name of the LAN

Event.

Saved in instrument state

Uppercase, Lowercase, Numeric, Symbol

Prior to A.02.00

Configure (Remote Command Only)

Allows the configuration of some of the above parameters from a single SCPI command.

Remote Command:

Example:

Initial S/W Revision:

:TRIGger[:SEQuence]:LXI:LAN[:SET]:CONFigure "lanEvent",

<enable>, <detection>, <delay>,<filter>,<identifier>

:TRIG:LXI:LAN:CONF "LAN0",1,FALL,0.0,"ALL","debugIdentifier"

Prior to A.02.00

Alarm

Pressing this key when it is not selected selects the alarm system as the LXI trigger. A new sweep/measurement starts when the configured IEEE 1588 time occurs if the LXI trigger is selected as the active trigger (see

“LXI Trigger” on page 321 ). Pressing this key when it is already selected accesses

the alarm source selection menu.

Key Path

Example

Preset

State Saved

Trigger, LXI Trigger

TRIG:ACP:SOUR ALAR

ON

Saved in instrument state

327

Common Measurement Functions 1

Status Bits/OPC dependencies The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event occurs and all the applicable trigger criteria have been met). A corresponding pop-up message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec.

This message goes away when a trigger signal appears.

Prior to A.02.00

Initial S/W Revision

Disable All

This key causes all Alarms in the trigger alarm list to go into the disabled state.

(Enabled = OFF)

Key Path

Remote Command

Example

Initial S/W Revision

Trigger, LXI Trigger, Alarm

:TRIGger[:SEQuence]:LXI:ALARm:DISable:ALL

:TRIG:LXI:ALAR:DIS:ALL

Prior to A.02.00

Alarm List

After selecting Alarm as the trigger source, the user is presented with a list of possible alarms. Pressing an alarm

(for example,. "ALARM0") branches to the alarm setup menu.

Key Path

Remote Command

Example

Preset

State Saved

Readback

Trigger, LXI Trigger, Alarm

:TRIGger[:SEQuence]:LXI:ALARm:LIST?

:TRIG:LXI:ALAR:LIST?

Returns the complete list of Alarm events which is: "ALARM0"

"ALARM0"

Saved in instrument state

Displays the value of the LXI Trigger Alarm Enabled parameter

(Enabled|Disabled).

Prior to A.02.00

Initial S/W Revision

Date/Time

Absolute alarm time sets an alarm for one specific time using the date and time of day (for example, 12/14/2007 at

11:45:15.3456). The Date and Time are represented in the instrument’s local time. This is the only way to set an alarm from the front panel.

Epoch time is another type of absolute alarm time. A specific time is identified by the number of seconds it occurs after January 1, 1970 00:00:00 in International Atomic Time (TAI). Epoch Time is time zone invariant. Epoch time

is only set via remote; see “Epoch Time Value (Remote Command Only)” on page 331 .

The date and time the alarm is scheduled to go off is noted on the branch softkey.

328

NOTE

Key Path

Readback

Initial S/W Revision

Key Path

Remote Command

Example

Notes

Preset

State Saved

Initial S/W Revision

Key Path

Remote Command

Common Measurement Functions 1

The Epoch Time Second and Epoch Time Fraction are the ultimate source of alarm information. The Absolute Time and Date may be changed from the front panel without being applied. When querying the Absolute Time and Date parameters from SCPI, if the

Absolute Time and Date have not been applied (and therefore do not match the Epoch

Time Second and Epoch Time Fraction), the string "(epoch time not set)" is added to the return value.

Trigger, LXI Trigger, Alarm, <alarmEvent>

Annotated with the date and time the alarm is scheduled to go off.

Prior to A.02.00

Trigger, LXI Trigger, Alarm, <alarmEvent>,Time

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]:ABSolu te "alarmEvent","date","time"

:TRIG:LXI:ALAR:TIME:ABS "ALARM0","2007/4/6", "15:45:02.123456"

"date" is a representation of the date the alarm should occur in the form of

"YYYY/MM/DD" where:

YYYY is the four digit representation of year. (for example, 2007)

MM is the two digit representation of month. (for example. 01 to 12)

DD is the two digit representation of day. (for example, 01 to 28, 29, 30, or 31 depending on the month and year)

"time" is a representation of the time of day the alarm should occur in the form of "HH:MM:SS.SSSSSS" where:

HH is the two digit representation of the hour in 24 hour format

MM is the two digit representation of minute

SS.SSSSSS is a real representing seconds (for example 02.123456)

Current date at initialization at 00:00:00.000000

Saved in instrument state

Prior to A.02.00

Trigger, LXI Trigger, Alarm, <alarmEvent>,Time

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]:ABSolu te? "alarmEvent"

329

Common Measurement Functions 1

Example

Notes

Preset

State Saved

Initial S/W Revision

:TRIG:LXI:ALAR:TIME:ABS? "ALARM0"

This query returns data using the following format "YYYY/MM/DD

HH:MM:SS.SSSSSS" If the Absolute time has been changed from the front panel, but has not been applied, the return value is of the form

"YYYY/MM/DD HH:MM:SS.SSSSSS (epoch time not set)".

<date> is a representation of the date the alarm should occur in the form of

YYYY/MM/DD where:

YYYY is the four digit representation of year. (for example, 2007)

MM is the two digit representation of month. (for example. 01 to 12)

DD is the two digit representation of day. (for example, 01 to 28, 29, 30, or 31 depending on the month and year)

<time> is a representation of the time of day the alarm should occur in the form of HH:MM:SS.SSSSSS where:

HH is the two digit representation of the hour in 24 hour format

MM is the two digit representation of minute

SS.SSSSSS is a real representing seconds (for example 02.123456)

Current date at initialization at 00:00:00.000000

Saved in instrument state

Prior to A.02.00

Date

The date the alarm should occur. All absolute alarm time parameters are set using the same SCPI command; however they each have their own front panel control.

When setting alarm values from the front panel, the new alarm time is not registered with the alarm system until the

"Set" key is pressed.

Key Path

Preset

State Saved

Range

Trigger, LXI Trigger, Alarm, <alarmEvent>,Time

Current date

Saved in instrument state current date – 214748/12/31. Values representing a time in the past result in an error.

Prior to A.02.00

Initial S/W Revision

Time

The time of the day, in the instrument’s local time (this takes into account time zones and daylight savings time), the alarm should occur. This parameter is based on a 24 hour clock.

All absolute alarm time parameters are set using the same SCPI command; however they each have their own front panel control.

When setting alarm values from the front panel, the new alarm time is not registered with the alarm system until the

330

Common Measurement Functions 1

"Set" key is pressed.

Key Path

Notes

Preset

State Saved

Range

Initial S/W Revision

Trigger, LXI Trigger, Alarm, <alarmEvent>,Time

Uses a 24 hour clock.

Values representing a time in the past result in an error.

Only valid time values are accepted.

The <second> field accepts a decimal number, and is valid to the microsecond position.

The <year>, <month>, <hour>, and <minute> fields all accept integers.

00:00:00.000000

Saved in instrument state

00:00:00.000000 – 23:59:59.999999

Prior to A.02.00

Apply

Causes the Absolute Alarm Time values to be converted into an Epoch time (see “Epoch Time Value (Remote

Command Only)” on page 331

), compared to the current time, and sent to the Alarm Trigger subsystem. This key can only be pressed when the epoch time and the absolute time are out of synch.

Key Path

Notes

Initial S/W Revision

Trigger, LXI Trigger, Alarm, <alarmEvent>,Time

Alarm times are settable to microsecond resolution.

Prior to A.02.00

Epoch Time Value (Remote Command Only)

Sets the LXI Alarm Time. This represents the number of seconds after January 1, 1970 00:00:00, in TAI time, that the alarm should go off.

Remote Command:

Example:

Notes:

Preset:

State Saved:

Range:

Initial S/W Revision:

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]

"alarmEvent",<seconds>, <fractionalSeconds>

:TRIG:LXI:ALAR:TIME "ALARM0",123456.0 S, 0.123456

Values representing a time in the past result in an error.

Seconds: The number of whole seconds between Jan 1, 1970 at 00:00:00 (in

TAI time) and the current date at initialization at 00:00:00 (in TAI time)

FractionalSeconds: 0

Saved in instrument state

Seconds: Epoch time of current date at 00:00:00 (in TAI time) –

253402300800 + number of seconds local time zone offset from UTC

FractionalSeconds: 0.0 – 0.999999

Prior to A.02.00

331

Common Measurement Functions 1

Remote Command:

Example:

Notes:

Preset:

State Saved:

Range:

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]?

:TRIG:LXI:ALAR:TIME?

Values representing a time in the past result in an error.

Seconds: The number of whole seconds between Jan 1, 1970 at 00:00:00 (in

TAI time) and the current date at initialization at 00:00:00 (in TAI time)

FractionalSeconds: 0

Saved in instrument state

Seconds: Epoch time of current date at 00:00:00 (in TAI time) –

253402300800 + number of seconds local time zone offset from UTC

FractionalSeconds: 0.0 – 0.999999

Prior to A.02.00

Initial S/W Revision:

Epoch Time Seconds (Remote Command Only)

Sets the seconds portion of the LXI Alarm time. This represents the number of seconds after January 1, 1970

00:00:00 (in TAI time) that the alarm should go off.

Values must be in the form of whole seconds; decimal values result in an error.

Remote Command:

Example:

Notes:

Preset:

State Saved:

Range:

Initial S/W Revision:

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME:SEConds

"alarmEvent",<seconds>

:TRIG:LXI:ALAR:TIME:SEC "ALARM0",123456.0 S

Values representing a time in the past result in an error.

Values containing a decimal portion result in an error.

The number of whole seconds between Jan 1, 1970 at 00:00:00 (in TAI time) and the current date at initialization at 00:00:00 (in TAI time)

Saved in instrument state

Epoch time of current date at 00:00:00 (in TAI time) – 253402300800 + number of seconds local time zone offset from UTC

Prior to A.02.00

Remote Command:

Example:

Notes:

Preset:

State Saved:

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME:SEConds?

"alarmEvent"

:TRIG:LXI:ALAR:TIME:SEC "ALARM0"?

Values representing a time in the past result in an error.

Values containing a decimal portion result in an error.

The number of seconds between Jan 1, 1970 at 00:00:00 (in TAI time) and the current date at initialization at 00:00:00 (in TAI time)

Saved in instrument state

332

Common Measurement Functions 1

Range: Epoch time of current date at 00:00:00 (in TAI time) – 253402300800 + number of seconds local time zone offset from UTC

Prior to A.02.00

Initial S/W Revision:

Epoch Time Fraction (Remote Command Only)

Sets the sub-second value of the Epoch time.

Remote Command:

Example:

Notes:

Preset:

State Saved:

Range:

Initial S/W Revision:

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]:FRACti on "alarmEvent",<fractionalSeconds>

:TRIG:LXI:ALAR:TIME:FRAC "ALARM0",0.123456 S

Values representing a time in the past result in an error.

0

Saved in instrument state

0.0 – 0.999999

Prior to A.02.00

Remote Command:

Example:

Notes:

Preset:

State Saved:

Min:

Max:

Initial S/W Revision:

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]:FRACti on? "alarmEvent"

:TRIG:LXI:ALAR:TIME:FRAC "ALARM0"?

Values representing a time in the past result in an error.

0

Saved in instrument state

0.0

0.999999

Prior to A.02.00

Relative Time (Remote Command Only)

Sets the values of Epoch Time Seconds and Epoch Time Fraction by adding an offset to the time when the command is issued. For example, if the Relative Time command is issued with an argument of 60s, the alarm will occur 1 minute in the future.

Remote Command:

Example:

Range:

Initial S/W Revision:

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]:RELati ve "alarmEvent",<seconds>

:TRIG:LXI:ALAR:TIME:REL "ALARM0",60.0s

0.0 – 1.7976931348623157 x 10308 (Max Double)

Prior to A.02.00

333

Common Measurement Functions 1

Remote Command:

Example:

Range:

Initial S/W Revision:

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:TIME[:VALue]:RELati ve? "alarmEvent"

:TRIG:LXI:ALAR:TIME:REL "ALARM0"?

0.0 – 1.7976931348623157 x 10308 (Max Double)

Prior to A.02.00

Period

Sets the amount of time that should elapse between alarms in a repeating alarm trigger.

Key Path

Remote Command

Example

Notes

Preset

State Saved

Range

Initial S/W Revision

Trigger, LXI Trigger, Alarm, <alarmEvent>

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:PERiod

"alarmEvent",<seconds>

:TRIG:LXI:ALAR:PER "ALARM0",1.2345 s

A period of 0.0s effectively causes the trigger to occur only once, since all repetitions are fired simultaneously

0.0 s

Saved in instrument state

0.0 – 1.7976931348623157 x 10308 (Max Double)

Prior to A.02.00

Key Path

Remote Command

Example

Notes

Preset

State Saved

Range

Initial S/W Revision

Trigger, LXI Trigger, Alarm, <alarmEvent>

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:PERiod?

"alarmEvent"

:TRIG:LXI:ALAR:PER "ALARM0"?

A period of 0.0s effectively causes the trigger to occur only once, since all repetitions are fired simultaneously

0.0 s

Saved in instrument state

0.0 – 1.7976931348623157 x 10308 (Max Double)

Prior to A.02.00

Repetitions

Sets the number of times a repeating alarm should fire once the initial alarm time has occurred.

Key Path

Remote Command

Trigger, LXI Trigger, Alarm, <alarmEvent>

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:REPeat

"alarmEvent",<repetitions>

334

Common Measurement Functions 1

Example

Notes

Preset

State Saved

Range

Initial S/W Revision

:TRIG:LXI:ALAR:REP "ALARM0",10

A repetition value of 0 means infinite repetitions (zero is a special case, triggers at the given period indefinitely)

A repetition value of 1 means 1 trigger only, not the intial trigger + 1 repeat

1

Saved in instrument state

1 – 2,147,483,647

Prior to A.02.00

Key Path

Remote Command

Example

Preset

State Saved

Min

Max

Initial S/W Revision

Trigger, LXI Trigger, Alarm, <alarmEvent>

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:REPeat?

"alarmEvent"

:TRIG:LXI:ALAR:REP "ALARM0",10

1

Saved in instrument state

1

2,147,483,647

Prior to A.02.00

Enabled

If Enabled is set to ON and the trigger source is set to ALARm, this alarm causes the instrument to trigger.

If Enabled is set to OFF, this alarm is ignored

Key Path

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Trigger, LXI Trigger, Alarm, <alarmEvent>

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:ENABled

"alarmEvent",ON|OFF|1|0

:TRIG:LXI:ALAR:ENAB "ALARM0",ON

OFF

Saved in instrument state

1|0

Prior to A.02.00

Key Path

Remote Command

Example

Trigger, LXI Trigger, Alarm, <alarmEvent>

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:ENABled?

"alarmEvent"

:TRIG:LXI:ALAR:ENAB "ALARM0"?

335

Common Measurement Functions 1

Preset

State Saved

Range

Initial S/W Revision

OFF

Saved in instrument state

1|0

Prior to A.02.00

Configure (Remote Command Only)

Allows the configuration of some of the above parameters from a single SCPI command.

Remote Command:

Example:

Initial S/W Revision:

:TRIGger[:SEQuence]:LXI:ALARm[:SET]:CONFigure

"alarmEvent", <enable>, <epochSeconds>, <epochFraction>,

<period>, <repeat>

:TRIG:LXI:ALAR:CONF "ALARM0",1,1000000.0,0.123456,1.2,3

Prior to A.02.00

Count (Remote Command Only)

Returns the number of alarms in the LXI Trigger Alarm List.

Remote Command:

Example:

Initial S/W Revision:

:TRIGger1|TRIGger[:SEQuence]:LXI:ALARm:COUNt?

:TRIG:LXI:ALAR:COUN?

Prior to A.02.00

Baseband I/Q

Pressing this key when it is not selected selects Baseband I/Q as the trigger. Pressing the key when it is already selected accesses the Baseband I/Q trigger type selection menu. The key is annotated to display which of the Baseband I/Q trigger types is currently selected.

Key Path

State Saved

Readback

Trigger

Saved in instrument state

The Baseband I/Q trigger source that becomes active when this key is selected is displayed. The possible values are "I/Q Mag", "I", "Q", "Input I", "Input Q", and "Aux I/Q Mag".

Prior to A.02.00

Initial S/W Revision

I/Q Mag

Pressing this key, when it is not selected, selects the I/Q magnitude signal as the trigger. The I/Q

Magnitude trigger condition is met when the I/Q magnitude crosses the I/Q magnitude trigger level. The magnitude is measured at the output of the main I/Q digital receiver.

Key Path

Example

Trigger, Baseband I/Q

TRIG:<meas>:SOUR IQM

336

Common Measurement Functions 1

Readback Text

Initial S/W Revision

I/Q Mag

Prior to A.02.00

Trigger Level

Sets a level for the I/Q magnitude trigger. When the signal crosses this level, with the chosen slope, the trigger occurs. If the specific Measurement displays the signal from the chosen sampling point a green line will be displayed to indicate the trigger level.

Key Path

Remote Command

Example

Notes

Preset

State Saved

Range

Readback Text

Initial S/W Revision

Trigger, Baseband I/Q, I/Q Mag

:TRIGger[:SEQuence]:IQMag:LEVel <ampl >

:TRIGger[:SEQuence]:IQMag:LEVel?

TRIG:IQM:LEV –30 dBm

The I/Q reference impedance is used for converting between power and voltage.

–25 dBm

Saved in instrument state

–200 dBm to 100 dBm

<level> dBm

Prior to A.02.00

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a falling edge.

Key Path

Remote Command

Example

Preset

State Saved

Initial S/W Revision

Trigger, Baseband I/Q, I/Q Mag

:TRIGger[:SEQuence]:IQMag:SLOPe POSitive | NEGative

:TRIGger[:SEQuence]:IQMag:SLOPe?

TRIG:IQM:SLOP POS

POSitive

Saved in instrument state

Prior to A.02.00

Trig Delay

Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT.

Key Path

Trigger, Baseband I/Q, I/Q Mag

337

Common Measurement Functions 1

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

:TRIGger[:SEQuence]:IQMag:DELay <time>

:TRIGger[:SEQuence]:IQMag:DELay?

:TRIGger[:SEQuence]:IQMag:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:IQMag:DELay:STATe?

TRIG:IQM:DEL 10 ms

TRIG:IQM:DEL:STAT ON

1 us

OFF

Saved in instrument state

–2.5 s to +10 s

Prior to A.02.00

I (Demodulated)

Pressing this key, when it is not selected, selects the main receiver's output I voltage as the trigger. The I

(Demodulated) trigger condition is met when the I voltage crosses the I voltage trigger level.

Key Path

Example

Readback Text

Initial S/W Revision

Trigger, Baseband I/Q

TRIG:<meas>:SOUR IDEM

I

Prior to A.02.00

Trigger Level

Sets a level for the I (Demodulated) trigger. When the signal crosses this level, with the chosen slope, the trigger occurs. If the specific Measurement displays the signal from the chosen sampling point a green line will be displayed to indicate the trigger level.

Key Path

Remote Command

Example

Preset

State Saved

Range

Readback Text

Initial S/W Revision

Trigger, Baseband I/Q, I (Demodulated)

:TRIGger[:SEQuence]:IDEMod:LEVel <voltage>

:TRIGger[:SEQuence]:IDEMod:LEVel?

TRIG:IDEM:LEV 0.5 V

0.25 V

Saved in instrument state

–1 to 1 V

0.1 of displayed unit (V, mV, etc.)

Prior to A.02.00

338

Common Measurement Functions 1

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a falling edge.

Key Path

Remote Command

Example

Preset

State Saved

Initial S/W Revision

Trigger, Baseband I/Q, I (Demodulated)

:TRIGger[:SEQuence]:IDEMod:SLOPe POSitive | NEGative

:TRIGger[:SEQuence]:IDEMod:SLOPe?

TRIG:IDEM:SLOP POS

POSitive

Saved in instrument state

Prior to A.02.00

Trig Delay

Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT.

Key Path

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Trigger, Baseband I/Q, I (Demodulated)

:TRIGger[:SEQuence]:IDEMod:DELay <time>

:TRIGger[:SEQuence]:IDEMod:DELay?

:TRIGger[:SEQuence]:IDEMod:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:IDEMod:DELay:STATe?

TRIG:IDEM:DEL 10 ms

TRIG:IDEM:DEL:STAT ON

1 us

OFF

Saved in instrument state

–2.5 s to +10 s

Prior to A.02.00

Q (Demodulated)

Pressing this key, when it is not selected, selects the main receiver's output Q voltage as the trigger. The

Q (Demodulated) trigger condition is met when the Q voltage crosses the Q voltage trigger level.

Key Path

Example

Readback Text

Initial S/W Revision

Trigger, Baseband I/Q

TRIG:<meas>:SOUR QDEM

Q

Prior to A.02.00

339

Common Measurement Functions 1

Trigger Level

Sets a level for the Q (Demodulated) trigger. When the signal crosses this level, with the chosen slope, the trigger occurs. If the specific Measurement displays the signal from the chosen sampling point a green line will be displayed to indicate the trigger level.

Key Path

Remote Command

Example

Preset

State Saved

Range

Readback Text

Initial S/W Revision

Trigger, Baseband I/Q, Q (Demodulated)

:TRIGger[:SEQuence]:QDEMod:LEVel <voltage>

:TRIGger[:SEQuence]:QDEMod:LEVel?

TRIG:QDEM:LEV 0.5 V

0.25 V

Saved in instrument state

–1 to 1 V

0.1 of displayed unit (V, mV, etc.)

Prior to A.02.00

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a falling edge.

Key Path

Remote Command

Example

Preset

State Saved

Initial S/W Revision

Trigger, Baseband I/Q, Q (Demodulated)

:TRIGger[:SEQuence]:QDEMod:SLOPe POSitive | NEGative

:TRIGger[:SEQuence]:QDEMod:SLOPe?

TRIG:QDEM:SLOP POS

POSitive

Saved in instrument state

Prior to A.02.00

Trig Delay

Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT.

Key Path

Remote Command

Example

Trigger, Baseband I/Q, Q (Demodulated)

:TRIGger[:SEQuence]:QDEMod:DELay <time>

:TRIGger[:SEQuence]:QDEMod:DELay?

:TRIGger[:SEQuence]:QDEMod:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:QDEMod:DELay:STATe?

TRIG:QDEM:DEL 10 ms

TRIG:QDEM:DEL:STAT ON

340

Common Measurement Functions 1

Preset

State Saved

Range

Initial S/W Revision

1 us

OFF

Saved in instrument state

–2.5 s to +10 s

Prior to A.02.00

Input I

Pressing this key, when it is not selected, selects the I channel's ADC voltage as the trigger. The Input I trigger condition is met when the voltage crosses the trigger level.

Key Path

Example

Readback Text

Initial S/W Revision

Trigger, Baseband I/Q

TRIG:<meas>:SOUR IINP

Input I

Prior to A.02.00

Trigger Level

Sets a level for the Input I trigger. When the signal crosses this level, with the chosen slope, the trigger occurs.

Key Path

Remote Command

Example

Preset

State Saved

Range

Readback Text

Initial S/W Revision

Trigger, Baseband I/Q, Input I

:TRIGger[:SEQuence]:IINPut:LEVel <voltage>

:TRIGger[:SEQuence]:IINPut:LEVel?

TRIG:IINP:LEV 0.5 V

0.25 V

Saved in instrument state

–1 to 1 V

0.1 of displayed unit (V, mV, etc.)

Prior to A.02.00

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a falling edge.

Key Path

Remote Command

Example

Preset

Trigger, Baseband I/Q, Input I

:TRIGger[:SEQuence]:IINPut:SLOPe POSitive | NEGative

:TRIGger[:SEQuence]:IINPut:SLOPe?

TRIG:IINP:SLOP POS

POSitive

341

Common Measurement Functions 1

State Saved

Initial S/W Revision

Saved in instrument state

Prior to A.02.00

Trig Delay

Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT.

Key Path

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Trigger, Baseband I/Q, Input I

:TRIGger[:SEQuence]:IINPut:DELay <time>

:TRIGger[:SEQuence]:IINPut:DELay?

:TRIGger[:SEQuence]:IINPut:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:IINPut:DELay:STATe?

TRIG:IINP:DEL 10 ms

TRIG:IINP:DEL:STAT ON

1 us

OFF

Saved in instrument state

–2.5 s to +10 s

Prior to A.02.00

Input Q

Pressing this key, when it is not selected, selects the Q channel's ADC voltage as the trigger. The Input Q trigger condition is met when the voltage crosses the trigger level.

Key Path

Example

Readback Text

Initial S/W Revision

Trigger, Baseband I/Q

TRIG:<meas>:SOUR QINP

Input Q

Prior to A.02.00

Trigger Level

Sets a level for the Input Q trigger. When the signal crosses this level, with the chosen slope, the trigger occurs.

Key Path

Remote Command

Example

Preset

Trigger, Baseband I/Q, Input Q

:TRIGger[:SEQuence]:QINPut:LEVel <voltage>

:TRIGger[:SEQuence]:QINPut:LEVel?

TRIG:QINP:LEV 0.5 V

0.25 V

342

Common Measurement Functions 1

State Saved

Range

Readback Text

Initial S/W Revision

Saved in instrument state

–1 to 1 V

0.1 of displayed unit (V, mV, etc.)

Prior to A.02.00

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a falling edge.

Key Path

Remote Command

Example

Preset

State Saved

Initial S/W Revision

Trigger, Baseband I/Q, Input Q

:TRIGger[:SEQuence]:QINPut:SLOPe POSitive | NEGative

:TRIGger[:SEQuence]:QINPut:SLOPe?

TRIG:QINP:SLOP POS

POSitive

Saved in instrument state

Prior to A.02.00

Trig Delay

Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT.

Key Path

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Trigger, Baseband I/Q, Input Q

:TRIGger[:SEQuence]:QINPut:DELay <time>

:TRIGger[:SEQuence]:QINPut:DELay?

:TRIGger[:SEQuence]:QINPut:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:QINPut:DELay:STATe?

TRIG:QINP:DEL 10 ms

TRIG:QINP:DEL:STAT ON

1 us

OFF

Saved in instrument state

–2.5 s to +10 s

Prior to A.02.00

Auxiliary Channel I/Q Mag

Pressing this key, when it is not selected, selects the Auxiliary Channel I/Q magnitude signal as the trigger. The Auxiliary Channel I/Q Magnitude trigger condition is met when the auxiliary receiver's I/Q

343

Common Measurement Functions 1

magnitude output crosses the Auxiliary I/Q magnitude trigger level.

Key Path

Example

Readback Text

Initial S/W Revision

Trigger, Baseband I/Q

TRIG:<meas>:SOUR AIQM

Aux I/Q Mag

Prior to A.02.00

Trigger Level

Sets a level for the I/Q magnitude trigger. When the signal crosses this level, with the chosen slope, the trigger occurs.

Key Path

Remote Command

Example

Notes

Preset

State Saved

Range

Readback Text

Initial S/W Revision

Trigger, Baseband I/Q, Aux Channel I/Q Mag

:TRIGger[:SEQuence]:AIQMag:LEVel <ampl >

:TRIGger[:SEQuence]:AIQMag:LEVel?

TRIG:AIQM:LEV –30 dBm

The I/Q reference impedance is used for converting between power and voltage.

–25 dBm

Saved in instrument state

–200 dBm to 100 dBm

<level> dBm

Prior to A.02.00

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a falling edge.

Key Path

Remote Command

Example

Preset

State Saved

Initial S/W Revision

Trigger, Baseband I/Q, Aux Channel I/Q Mag

:TRIGger[:SEQuence]:AIQMag:SLOPe POSitive | NEGative

:TRIGger[:SEQuence]:AIQMag:SLOPe?

TRIG:AIQM:SLOP POS

POSitive

Saved in instrument state

Prior to A.02.00

Trig Delay

Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger criteria. You

344

Common Measurement Functions 1

can use negative delay to pre-trigger the instrument in time domain or FFT.

Key Path

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Trigger, Baseband I/Q, Aux Channel I/Q Mag

:TRIGger[:SEQuence]:AIQMag:DELay <time>

:TRIGger[:SEQuence]:AIQMag:DELay?

:TRIGger[:SEQuence]:AIQMag:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:AIQMag:DELay:STATe?

TRIG:AIQM:DEL 10 ms

TRIG:AIQM:DEL:STAT ON

1 us

OFF

Saved in instrument state

–2.5 s to +10 s

Prior to A.02.00

Trigger Center Frequency

This key sets the center frequency to be used by the auxiliary receiver.

Key Path

Remote Command

Example

Notes

Preset

State Saved

Range

Initial S/W Revision

Trigger, Baseband I/Q, Aux Channel I/Q Mag

:TRIGger[:SEQuence]:AIQMag:CENTer <freq>

:TRIGger[:SEQuence]:AIQMag:CENTer?

:TRIG:AIQM:CENT 10 MHz

Trigger CF + 1/2 Trigger BW < Max

Trigger CF – 1/2 Trigger BW > Min

0 Hz

Saved in instrument state

–40 MHz to 40 MHz

Prior to A.02.00

Trigger Bandwidth

This key sets the information bandwidth used by the auxiliary receiver for the Auxiliary Channel I/Q Magnitude trigger.

Key Path

Remote Command

Trigger, Baseband I/Q, Aux Channel I/Q Mag

:TRIGger[:SEQuence]:AIQMag:BANDwidth <freq>

:TRIGger[:SEQuence]:AIQMag:BANDwidth?

345

Common Measurement Functions 1

Example

Notes

Preset

State Saved

Range

Initial S/W Revision

:TRIG:AIQM:BAND 8 MHz

The combined sample rate for the main and auxiliary receivers cannot exceed

100 MSa/sec. The bandwidth available to the Trigger BW is limited to what is available after the main receiver's bandwidth (Info BW, sometimes pre-FFT

BW) is set. Because of this limitation, the Max is not always achievable.

The combination of Trigger Center Freq and Trigger BW is also limited:

Trigger CF + 1/2 Trigger BW < Max

Trigger CF – 1/2 Trigger BW > Min

Bandwidth option dependent:

No Opt: 10 MHz

Opt B25: 25 MHz

Opt S40: 40 MHz

Saved in instrument state

10 Hz to Maximum

Prior to A.02.00

Auto/Holdoff

Opens up a menu that lets you adjust Auto Trigger and Trigger Holdoff parameters

Key Path

Readback line

Initial S/W Revision

Trigger

Displays a summaryof the Auto Trig and Holdoff settings, in square brackets

First line: Auto Off or Auto On

Second Line: “Hldf” followed by:

• If Holdoff is Off, readback Off

• If Holdoff On and Type = Normal, readback value

• If Holdoff On and Type = Above, readback value followed by AL

• If Holdoff On and Type = Below, readback value followed by BL

• If Holdoff Type selection is not supported by the current measurement,

Holdoff Type is always Normal

A.02.00

Auto Trig

Sets the time that the analyzer will wait for the trigger conditions to be met. If they are not met after that much time, then the analyzer is triggered anyway.

Key Path

Trigger, Auto/Holdoff

346

Common Measurement Functions 1

Remote Command

Example

Notes

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

:TRIGger[:SEQuence]:ATRigger <time>

:TRIGger[:SEQuence]:ATRigger?

:TRIGger[:SEQuence]:ATRigger:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:ATRigger:STATe?

TRIG:ATR:STAT ON

TRIG:ATR 100 ms

The "time that the analyzer will wait" starts when the analyzer is ready for a trigger, which may be hundreds of ms after the data acquisition for a sweep is done. The "time" ends when the trigger condition is satisfied, not when the delay ends.

Off, 100 ms

Saved in instrument state

1 ms

100 s s

Prior to A.02.00

Trig Holdoff

Sets the holdoff time between triggers. When the trigger condition is satisfied, the trigger occurs, the delay begins, and the holdoff time begins. New trigger conditions will be ignored until the holdoff time expires. For a free-running trigger, the holdoff value is the minimum time between triggers.

Key Path

Remote Command

Example

Preset

State Saved

Min

Max

Default Unit

Initial S/W Revision

Trigger, Auto/Holdoff

:TRIGger[:SEQuence]:HOLDoff <time>

:TRIGger[:SEQuence]:HOLDoff?

:TRIGger[:SEQuence]:HOLDoff:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:HOLDoff:STATe?

TRIG:HOLD:STAT ON

TRIG:HOLD 100 ms

Off, 100 ms

Saved in instrument state

0 s

0.5 s s

Prior to A.02.00

347

Common Measurement Functions 1

Holdoff Type

Lets you set the Trigger Holdoff Type.

NOTE

Holdoff Type is not supported by all measurements. If the current measurement does not support it, this key will be blank and the Holdoff Type will be Normal. If the Holdoff Type SCPI is sent while in such a measurement, the SCPI will be accepted and the setting remembered, but it will have no effect until a measurement is in force that supports Holdoff Type.

Trigger Holdoff Type functionality:

NORMal

This is the “oscilloscope” type of trigger holdoff, and is the setting when the Holdoff Type key does not appear. In this type of holdoff, no new trigger will be accepted until the holdoff interval has expired after the previous trigger.

ABOVe

If the trigger slope is positive, a trigger event is generated only if the signal characteristic of interest crosses the trigger threshold (with positive slope) and then remains above the threshold for at least the holdoff time. For negative slope, the trigger event is generated if the signal characteristic crosses the threshold (with negative slope) after having been above the threshold for at least the holdoff time. In either case, the trigger event is associated with the time the level was crossed.

BELow

If the trigger slope is positive, a trigger event is generated only if the signal characteristic of interest crosses the trigger threshold (with positive slope) after having been below the threshold for at least the holdoff time. For negative slope, the trigger event is generated if the signal characteristic crosses the threshold (with negative slope) and then remains below the threshold for at least the holdoff time. In either case, the trigger event is associated with the time the level was crossed.

Key Path

Remote Command

Example

Preset

State Saved

Initial S/W Revision

Trigger, Auto/Holdoff

:TRIGger[:SEQuence]:HOLDoff:TYPE NORMal|ABOVe|BELow

:TRIGger[:SEQuence]:HOLDoff:TYPE?

TRIG:HOLD:TYPE NORM

All modes but GSM/EDGE: Normal

GSM/EDGE: Below

Saved in instrument state

A.02.00

348

Common Measurement Functions 1

Trigger Offset (Remote Command Only)

ESA Backwards Compatibility command

Remote Command:

Example:

Notes:

Preset:

State Saved:

Min:

Max:

Initial S/W Revision:

:TRIGger[:SEQuence]:OFFSet <time>

:TRIGger[:SEQuence]:OFFSet?

:TRIGger[:SEQuence]:OFFSet:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:OFFSet:STATe?

TRIG:OFFS ON

TRIG:OFFS –100 ms

These are ESA commands for trigger offset that allowed you to use a positive or negative delay when in zero span and in a Res BW >= 1 kHz. For ESA compatibility, X-series analyzers keep track of this offset and add it to the

Trigger Delay for line, video or external whenever the value is sent to the hardware, if in Zero Span and RBW >= 1 kHz.

Off, 0 s

Saved in instrument state

–11 s

+11 s

Prior to A.02.00

View/Display

The View/Display key opens up the Display Menu (common to most measurments) and the View menu for the current measurement.

Some measurements have simple View menus, or even no View menu, others provide many different

Views.

Views are different ways of looking at data, usually different ways of looking at the same data, especially when the data represents a time record that is being digitally processed with an FFT and/or other digital signal processing algorithms.

Key Path

Initial S/W Revision

Front-panel key

Prior to A.02.00

Display

The

Display

menu is common to most measurements, and is used for configuring items on the display.

Some

Display

menu settings apply to all the measurements in a mode, and some only to the current measurement. Those under the

System Display Settings

key apply to all measurements in all modes.

Key Path

Display

349

Common Measurement Functions 1

Initial S/W Revision Prior to A.02.00

Annotation

Turns on and off various parts of the display annotation. The annotation is divided up into four categories:

1. Meas Bar: This is the measurement bar at the top of the screen. It does not include the settings panel or the Active Function. Turning off the Meas Bar turns off the settings panel and the Active

Function. When the Meas Bar is off, the graticule area expands to fill the area formerly occupied by the Meas Bar.

2. Screen Annotation: this is the annotation and annunciation around the graticule, including any annotation on lines (such as the display line, the threshold line, etc.) This does NOT include the marker number or the N dB result. When off, the graticule expands to fill the entire graticule area.

3. Trace annotation: these are the labels on the traces, showing their detector (or their math mode).

4. Active Function annotation: this is the active function display in the meas bar, and all of the active function values displayed on softkeys.

See the figure below. Each type of annotation can be turned on and off individually.

350

Common Measurement Functions 1

Key Path

Initial S/W Revision

View/Display, Display

Prior to A.02.00

Meas Bar On/Off

This function turns the Measurement Bar on and off, including the settings panel. When off, the graticule area expands to fill the area formerly occupied by the Measurement Bar.

Remote Command:

Example:

Dependencies:

Preset:

:DISPlay:ANNotation:MBAR[:STATe] OFF|ON|0|1

:DISPlay:ANNotation:MBAR[:STATe]?

DISP:ANN:MBAR OFF

Grayed out and forced to OFF when

System Display Settings

,

Annotation

is set to Off.

On

This should remain Off through a Preset when

System Display Settings

,

Annotation

is set to Off.

351

Common Measurement Functions 1

State Saved:

Key Path:

Initial S/W Revision:

Saved in instrument state.

View/Display, Display, Annotation

Prior to A.02.00

Screen

This controls the display of the annunciation and annotation around the graticule, including any annotation on lines

(such as the display line, the threshold line, etc.) and the y-axis annotation. This does NOT include marker annotation (or the N dB result). When off, the graticule expands to fill the entire graticule area, leaving only the

1.5% gap above the graticule as described in the Trace/Detector chapter.

Remote Command:

Example:

Dependencies:

Preset:

State Saved:

Key Path:

Initial S/W Revision:

:DISPlay:ANNotation:SCReen[:STATe] OFF|ON|0|1

:DISPlay:ANNotation:SCReen[:STATe]?

DISP:ANN:SCR OFF

Grayed out and forced to OFF when

System Display Settings

,

Annotation

is set to Off.

On

This should remain Off through a Preset when

System Display Settings

,

Annotation

is set to Off

Saved in instrument state.

View/Display, Display, Annotation

Prior to A.02.00

Trace

Turns on and off the labels on the traces, showing their detector (or their math mode) as described in the

Trace/Detector section.

If trace math is being performed with a trace, then the trace math annotation will replace the detector annotation.

Remote Command:

Example:

Preset:

State Saved:

Key Path:

Initial S/W Revision:

:DISPlay:ANNotation:TRACe[:STATe] ON|OFF|1|0

:DISPlay:ANNotation:TRACe[:STATe]?

DISP:ANN:TRAC OFF

Off

Saved in instrument state.

View/Display, Display, Annotation

Prior to A.02.00

Active Function Values On/Off

Turns on and off the active function display in the Meas Bar, and all of the active function values displayed on the softkeys.

352

Common Measurement Functions 1

Note that all of the softkeys that have active functions have these numeric values blanked when this function is on.

This is a security feature..

Remote Command:

Example:

Dependencies:

Preset:

State Saved:

Key Path:

Initial S/W Revision:

:DISPlay:ACTivefunc[:STATe] ON|OFF|1|0

:DISPlay:ACTivefunc[:STATe]?

DISP:ACT OFF

Grayed out and forced to OFF when

System Display Settings

,

Annotation

is set to Off.

On

This should remain Off through a Preset when

System Display Settings

,

Annotation

is set to Off

Saved in instrument state.

View/Display, Display, Annotation

Prior to A.02.00

Title

Displays menu keys that enable you to change or clear a title on your display.

Key Path

Initial S/W Revision

View/Display, Display

Prior to A.02.00

353

Common Measurement Functions 1

Change Title

Writes a title into the "measurement name" field in the banner, for example, "Swept SA".

Press Change Title to enter a new title through the alpha editor. Press Enter or Return to complete the entry. Press

ESC to cancel the entry and preserve your existing title.

The display title will replace the measurement name. It remains for this measurement until you press

Change Title

again, or you recall a state, or a Preset is performed. A title can also be cleared by pressing

Title

,

Clear Title

.

NOTE

Notice the inclusion of the <measurement> parameter in the command below. Because each measurement remembers the Display Title, the command must be qualified with the measurement name. For the Swept SA measurement this is not the case; for backwards compatibility, no <measurement> parameter is used when changing the Display Title for the Swept SA measurement.

Remote Command

Example

Initial S/W Revision

Key Path

Mode

Notes

Preset

State Saved

:DISPlay:<measurement>:ANNotation:TITLe:DATA <string>

:DISPlay:<measurement>:ANNotation:TITLe:DATA?

DISP:ANN:TITL:DATA "This Is My Title"

This example is for the Swept SA measurement in the Spectrum Analyzer mode. The SANalyzer <measurement> name is not used.

DISP:ACP:ANN:TITL:DATA "This Is My Title"

This example is for Measurements other than Swept SA.

Both set the title to: This Is My Title

Prior to A.02.00

View/Display, Display, Title

All

Pressing this key cancels any active function.

When a title is edited the previous title remains intact (it is not cleared) and the cursor goes at the end so that characters can be added or BKSP can be used to go back over previous characters.

No title (measurement name instead)

Saved in instrument state.

Clear Title

Clears a title from the front-panel display. Once cleared, the title cannot be retrieved. After the title is cleared, the current Measurement Name replaces it in the title bar.

Example:

Notes:

DISP:ANN:TITL:DATA "" clears any existing title characters.

Use the :DISPlay:ANNotation:TITLe:DATA <string> command with an empty string.

354

Common Measurement Functions 1

Preset:

Key Path:

Initial S/W Revision:

Performed on Preset.

View/Display, Display, Title

Prior to A.02.00

Graticule

Pressing Graticule turns the display graticule On or Off. It also turns the graticule y-axis annotation on and off.

Remote Command:

Example:

Notes:

Preset:

State Saved:

Key Path:

Initial S/W Revision:

:DISPlay:WINDow[1]:TRACe:GRATicule:GRID[:STATe]

OFF|ON|0|1

:DISPlay:WINDow[1]:TRACe:GRATicule:GRID[:STATe]?

DISP:WIND:TRAC:GRAT:GRID OFF

The graticule is the set of horizontal and vertical lines that make up the grid/divisions for the x-axis and y-axis.

On

Saved in instrument state

View/Display, Display

Prior to A.02.00

Display Line

Activates an adjustable horizontal line that is used as a visual reference line. The line’s vertical position corresponds to its amplitude value. The value of the display line (for example, "–20.3 dBm") appears above the line itself on the right side of the display in the appropriate font.

The display line can be adjusted using the step keys, knob, or numeric keypad. The unit of the Display

Line is determined by the Y axis unit setting under Amplitude. If more than one window has a display line, the display line of the selected window is controlled.

If the display line is off the screen, it shows as a line at the top/bottom of the screen with an arrow pointing up or down. As with all such lines (Pk Thresh, Trigger Level, etc.) it is drawn on top of all traces.

The display line is unaffected by Auto Couple.

Remote Command:

Example:

:DISPlay:WINDow[1]:TRACe:Y:DLINe <ampl>

:DISPlay:WINDow[1]:TRACe:Y:DLINe?

:DISPlay:WINDow[1]:TRACe:Y:DLINe:STATe OFF|ON|0|1

:DISPlay:WINDow[1]:TRACe:Y:DLINe:STATe?

DISP:WIND:TRAC:Y:DLIN:STAT ON

DISP:WIND:TRAC:Y:DLIN:STAT –32 dBm

355

Common Measurement Functions 1

Preset: Set the Display Line to Off and –25 dBm on Preset. When the Display Line goes from Off to On, if it is off screen, set it to either the top or bottom of screen, depending on which direction off screen it was.

The Display Line's value does not change when it is turned off.

State Saved:

Min:

Max:

Key Path:

Default Unit:

Initial S/W Revision:

Saved in instrument state.

∞ (minus infinity) in current units

+

∞ (plus infinity) in current units

View/Display, Display

Depends on the current selected Y axis unit

Prior to A.02.00

System Display Settings

These settings are "Mode Global" – they affect all modes and measurements and are reset only by

Restore Misc Defaults

or

Restore System Defaults

under System.

Key Path

Initial S/W Revision

View/Display, Display

Prior to A.02.00

Annotation Local Settings

This is a Mode Global override of the meas local annotation settings. When it is

All Off

, it forces

Screen

Annotation, Meas Bar, Trace, and Active Function Values

settings to be

OFF

for all measurements in all modes.

This provides the security based "annotation off" function of previous analyzers; hence it uses the legacy SCPI command.

When it is

All Off,

the

Screen, Meas Bar, Trace,

and

Active Function Values

keys under the

Display, Annotation

menu are grayed out and forced to

Off

. When

Local Settings

is selected, you are able to set the local annotation settings on a measurement by measurement basis.

Remote Command:

Example:

Preset:

State Saved:

Key Path:

Initial S/W Revision:

:DISPlay:WINDow[1]:ANNotation[:ALL] OFF|ON|0|1

:DISPlay:WINDow[1]:ANNotation[:ALL]?

:DISP:WIND:ANN OFF

On (Set by Restore Misc Defaults)

Not saved in instrument state.

View/Display, Display, System Display Settings

Prior to A.02.00

Theme

This key allows you to change the Display theme. This is similar to the Themes selection under Page Setup and

356

Common Measurement Functions 1

Save Screen Image. The four themes are detailed below.

Remote Command:

Preset:

State Saved:

Key Path:

Notes:

Example:

Initial S/W Revision:

:DISPlay:THEMe TDColor|TDMonochrome|FCOLor|FMONochrome

:DISPlay:THEMe?

TDColor (Set by Restore Misc Defaults)

Not saved in instrument state.

View/Display, Display, System Display Settings

TDColor – 3D is the standard color theme with filling and shading

TDMonochrome – is similar to 3D color, but only black is used

FCOLor – flat color is intended for inkjet printers to conserve ink. It uses a white background instead of black.

FMONochrome – is like flat color, but only black is used

DISP:THEM TDM sets the display theme to 3D Monochrome.

Prior to A.02.00

Backlight

Accesses the display backlight on/off keys. This setting may interact with settings under the Windows "Power" menu.

When the backlight is off, pressing ESC, TAB, SPACE, ENTER, UP, DOWN, LEFT, RIGHT, DEL, BKSP, CTRL, or ALT turns the backlight on without affecting the application. Pressing any other key will turn backlight on and could potentially perform the action as well.

Remote Command:

Preset:

Key Path:

Initial S/W Revision:

:DISPlay:BACKlight ON|OFF

:DISPlay:BACKlight?

ON (Set by Restore Misc Defaults)

View/Display, Display, System Display Settings

Prior to A.02.00

On

Turns the display backlight on.

Example:

Key Path:

Readback:

Initial S/W Revision:

DISP:BACK ON

View/Display, Display, System Display Settings, Backlight

On

Prior to A.02.00

357

Common Measurement Functions 1

Off

Turns the display backlight off.

Example:

Key Path:

Readback:

Initial S/W Revision:

DISP:BACK OFF

View/Display, Display, System Display Settings, Backlight

Off

Prior to A.02.00

Backlight Intensity

An active function used to set the backlight intensity. It goes from 0 to 100 where 100 is full on and 0 is off. This value is independent of the values set under the Backlight on/off key.

Remote Command:

Example:

Preset:

Min:

Max:

Key Path:

Initial S/W Revision:

:DISPlay:BACKlight:INTensity <integer>

:DISPlay:BACKlight:INTensity?

DISP:BACK:INT 50

100 (Set by Restore Misc Defaults)

0

100

View/Display, Display, System Display Settings

Prior to A.02.00

Full Screen

When

Full Screen

is pressed the measurement window expands horizontally over the entire instrument display. The screen graticule area expands to fill the available display area.

It turns off the display of the softkey labels, however the menus and active functions still work. (Though it would obviously be very hard to navigate without the key labels displayed.) Pressing

Full Screen

again while Full Screen is in effect cancels Full Screen.

Note that the banner and status lines are unaffected. You can get even more screen area for your data display by turning off the Meas Bar (in the Display menu) which also turns off the settings panel.

Full Screen is a Meas Global function. Therefore it is cancelled by the

Preset

key.

Remote Command:

Preset:

State Saved:

Key Path:

:DISPlay:FSCReen[:STATe] OFF|ON|0|1

:DISPlay:FSCReen[:STATe]?

Off

Not saved in instrument state.

Display

358

Common Measurement Functions 1

Backwards Compatibility SCPI: :DISPlay:MENU[:STATe] OFF|ON|0|1

Backwards Compatibility SCPI: DISPlay:MENU[:STATe] emulates ESA full screen functionality, which is the same as the FSCReen command in PSA except that the sense of on/off is reversed (that is, OFF means the menus are OFF) and the default is ON.

Initial S/W Revision: Prior to A.02.00

Display Enable (Remote Command Only)

Turns the display on/off, including the display drive circuitry. The backlight stays lit so you can tell that the instrument is on. The display enable setting is mode global. The reasons for turning the display off are three:

• To increase speed as much as possible by freeing the instrument from having to update the display

• To reduce emissions from the display, drive circuitry

• For security purposes

If you have turned off the display:

• and you are in local operation, the display can be turned back on by pressing any key or by sending the SYSTem:DEFaults MISC command or the DISPlay:ENABle ON (neither *RST nor

SYSTem:PRESet enable the display.)

• and you are in remote operation, the display can be turned back on by pressing the

Local

or

Esc

keys or by sending the SYSTem:DEFaults MISC command or the DISPlay:ENABle ON (neither *RST nor SYSTem:PRESet enable the display.) and you are using either the SYSTem:KLOCk command or GPIB local lockout, then no front-panel key press will turn the display back on. You must turn it back on remotely.

Remote Command:

Example:

Couplings:

:DISPlay:ENABle OFF|ON|0|1

:DISPlay:ENABle?

DISP:ENAB OFF

DISP:ENAB OFF turns Backlight OFF and DISP:ENAB ON turns Backlight

ON. However, settings of Backlight do not change the state of DISP:ENAB

Preset:

State Saved:

Initial S/W Revision:

On

Set by SYST:DEF MISC, but Not affected by *RST or SYSTem:PRESet.

Not saved in instrument state.

Prior to A.02.00

359

Common Measurement Functions 1

360

Common Measurement Functions 2

Common Measurement Functions 2

The key and command descriptions in this section describe functions that operate identically in multiple measurements and/or modes. This section is a library of functions that is referenced by many measurements and modes

To find the exact description and parameters for functions in a specific measurement, always look in the measurement section of this documentation. Pressing the front-panel key or softkey and then pressing the green Help key also provides the correct information.

NOTE

If you want to print the documentation, be sure to select this section and the measurement of interest to ensure having all the information you need. See

“Printing Acrobat Files” on page 81

for further instructions about printing.

AMPTD Y Scale (Amplitude)

This menu has controls for the input signal conditioning as well as the Y-scaling of trace data. Input signal conditioning actually affect the input signal and the associated measurement quality, whereas

Y-scaling is non-destructive of data. Even if the data is scaled so as to be clipped or completely off the display, the marker readouts are still correct and accurate data may still be retrieved via SCPI.

Key Path

Initial S/W Revision

Front Panel

Prior to A.02.00

Range

The Range setting represents the amplitude of the largest sinusoidal signal that could be present within the IF without being clipped by the ADC. For signals with high peak-to-rms ratios, the range may need to exceed the rms signal power by a fair amount to avoid clipping.

Key Path

Mode

Remote Command

Example

Notes

Preset

State Saved

Min

AMPTD/YScale

VSA, LTE, LTETDD, IDEN

[:SENSe]:POWer[:RF]:RANGe <real>

[:SENSe]:POWer[:RF]:RANGe?

POW:RANG 25

POW:RANG?

The parameter is interpreted as dBm

20

Saved in instrument state.

depends on model and preamp options

361

Common Measurement Functions 2

Max

Initial S/W Revision

Modified at S/W Revision depends on model and preamp options

Prior to A.02.00

A.02.00

Y Axis Scaling

Y axis scaling allows you to view the entire range of the data or zoom in on a range of interest. Scaling does not affect measurement setup, and rescaling can be done at any time on paused or complete measurements and the results of the rescaling are immediately visible. Y scaling can be made to track range setting for convenience in setting up measurements.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

AMPTD Y-Scale

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Select Trace

This function is a duplicate of the same function found on the Trace/Detector menu. The Select Trace key is also located here to allow you to conveniently choose which trace the Y scaling applies.

See

“Select Trace” on page 428 for details.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

AMPTD Y-Scale, Y Axis Scaling

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Y Auto Scale

This immediate action key causes the Y reference value and Scale per Division to change so as to display the full trace without clipping.

Key Path

Mode

Measurement

Remote Command

Example

Initial S/W Revision

Modified at S/W Revision

AMPTD/YScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:AUTO:ONCE

:DISP:VECT:TRAC1:Y:AUTO:ONCE

Prior to A.02.00

A.02.00

362

Common Measurement Functions 2

Y Reference Value

This function controls the Y value of the selected trace at the Reference Position. It has no effect on hardware input settings.

See “Y Reference: Position” on page 364

for more details.

Key Path

Mode

Measurement

Remote Command

Example

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

AMPTD/YScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:RLEVel <real>

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:RLEVel?

DISP:VECT:TRAC:Y:RLEV 20

DISP:VECT:TRAC:Y:RLEV?

None. This does not affect any hardware input settings.

Depends on trace

Saved in instrument state.

–9.9E+37

9.9E+37

Prior to A.02.00

A.02.00

Y Scale Per Division

This controls the Y scale per division of the selected trace.

Key Path

Mode

Measurement

Remote Command

Example

Couplings

Preset

State Saved

AMPTD/YScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:PDIVision

<real>

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:PDIVision?

DISP:VECT:TRAC:Y:PDIV 10

DISP:VECT:TRAC:Y:PDIV?

None.

Depends on trace

Saved in instrument state.

363

Common Measurement Functions 2

Min

Max

Initial S/W Revision

Modified at S/W Revision

–9.9E+37

9.9E+37

Prior to A.02.00

A.02.00

Couple Ref to Range

When Couple Ref to Range is on, a Y scaling is adjusted when the Range changes. For example, on traces with Y units of dBm, the reference value changes by the same amount in dB as the Range does. On a trace with Y units of Volts, the Per Division setting changes by a factor of approx. 1.25 when the Range changes by 2 dB. This function may be turned on or off on for each individual trace.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

AMPTD/YScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:RLEVel:AUTO

OFF|ON|0|1

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:RLEVel:AUTO?

DISP:VECT:TRAC1:Y:RLEV:AUTO ON

DISP:VECT:TRAC1:Y:RLEV:AUTO?

Range coupling is not available for Phase and Group delay traces.

1

Saved in instrument state.

On | Off

Prior to A.02.00

A.02.00

Y Reference: Position

This determines the position of the reference line for Y scaling for the selected trace. It may be set to the top, bottom, or center of the grid.

Key Path

Mode

Measurement

AMPTD/YScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

364

Common Measurement Functions 2

Remote Command

Example

Couplings

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:RPOSition

TOP|CENTer|BOTTom

:DISPlay:<meas>:TRACe[1]|2|3|4:Y[:SCALe]:RPOSition?

DISP:VECT:TRAC1:Y:RPOS TOP

DISP:VECT:TRAC1:Y:RPOS?

Changing trace format or data can affect this. Each format "remembers" its reference position.

Depends on trace format and trace data. Top for LogMag or most LinearMag traces, middle for Real, Imaginary, Vector displays, Eye diagrams, Phase,

Delay, Bottom for Linear Mag EVM

Saved in instrument state.

Top|Ctr|Bottom

Prior to A.02.00

A.02.00

Y Unit Preference

This determines the preferred Y unit for the selected trace. You can select Peak, RMS, Power units, or an automatic selection. The automatic selection is to show Power units for frequency domain data and Peak units for time domain data.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Range

Readback Text

Initial S/W Revision

Modified at S/W Revision

AMPTD/YScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:Y:UNIT:PREFerence

AUTO|PEAK|RMS|POWer|MRMS

:DISPlay:<meas>:TRACe[1]|2|3|4:Y:UNIT:PREFerence?

DISP:VECT:TRAC1:Y:UNIT:PREF PEAK

DISP:VECT:TRAC1:Y:UNIT:PREF?

AUTO

Saved in instrument state.

AUTO|PEAK|RMS|POW|MRMS

Auto|Peak|RMS|Power|mRMS

Prior to A.02.00

A.02.00

The following SCPI only command can be used to determine exactly which Y unit was chosen based on

365

Common Measurement Functions 2

the setting of the above:

Key Path

Mode

Measurement

Remote Command

Example

Initial S/W Revision

Modified at S/W Revision

AMPTD/YScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:Y:UNIT?

DISP:VECT:TRAC1:Y:UNIT?

Prior to A.02.00

A.02.00

Y Log Ratio

This property is only used if the Trace Format is set to LogMag (Linear Unit). In this format type, you set the Y Log Ratio instead of Y Scale Per Division to determine Y scaling. It sets the ratio of the top of the

Y axis to the bottom.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

AMPTD/YScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:Y:LRATio <real>

:DISPlay:<meas>:TRACe[1]|2|3|4:Y:LRATio?

DISP:VECT:TRAC1:Y:LRAT 10000

DISP:VECT:TRAC1:Y:LRAT?

This is grayed out if the trace format is not Log Mag (linear unit).

100000

Saved in instrument state.

1.001

100e6

Prior to A.02.00

A.02.00

Vector Horiz Center

The Vector trace formats are I-Q and Constellation. When you are in one of these formats you set the vertical (imaginary) axis scaling with the Y Reference Value, Y Reference Position, and Y Scale Per

Division properties. The scaling of the horizontal axis is set so as to maintain an aspect ratio of 1:1. The

366

Common Measurement Functions 2

Vector Horiz Center property is used to set the position of the origin.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

AMPTD/YScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:VHCenter <real>

:DISPlay:<meas>:TRACe[1]|2|3|4:VHCenter?

DISP:DDEM:TRAC1:VHC 0.2

DISP:DDEM:TRAC1:VHC?

0

Saved in instrument state.

–9.9e37

9.9e37

Prior to A.02.00

A.02.00

Copy Y Scale

This front-panel only function copies the following Y scaling information from the selected trace to another:

• Y reference Position

• Y Reference Value

• Y Unit Preference

• Vector Horiz Center

• Couple Ref to Range

• Y Log Ratio

• Y Reference Line

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

AMPTD Y-Scale, Y Axis Scaling

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

367

Common Measurement Functions 2

Reference Line

This controls whether the Y reference line is visible or not.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Initial S/W Revision

Modified at S/W Revision

AMPTD/YScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:RLINe OFF|ON|0|1

:DISPlay:<meas>:TRACe[1]|2|3|4:RLINe?

DISP:VECT:TRAC1:RLIN ON

DISP:VECT:TRAC1:RLIN?

OFF

Saved in instrument state.

Prior to A.02.00

A.02.00

BW (Bandwidth)

The BW key allows you to control the resolution bandwidth of the spectrum measurement result, as well as the shape of the resolution bandwidth filter (controlled by the FFT windowing function).

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Front Panel

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Res BW

This key allows you to select the resolution bandwidth of the measurement. Res BW is mathematically related to Time length and Window type, so changing one of these, directly or indirectly, must change at least one other.

Res BW and Time length are related by the following equation:

Res BW = ENBW / T where:

ENBW is the normalized effective noise bandwidth of the Window. See “FFT Window” on page 371

for more details).

T is the time record length.

Therefore, if you change Res BW, Main Time must also change, and vice versa. (If the Gate function is on, then it is Gate Length, not Main Time, that is related to Res BW by the above equation.)

368

Common Measurement Functions 2

For convenience, Res BW is by default also coupled to Span (but not vice versa). This coupling may be turned off. See

“Res BW Coupling” on page 369 for more details.

Limits:

The minimum Res Bw to Span ratio is related to the maximum Main Time length, and is given by:

ENBW / 409600 if Freq points state parameter is set to Auto

ENBW / (Freq Points – 1)if Freq points parameter is manually set

The maximum Res BW to Span ratio is related to the minimum time record size (16 points for most windows, 17 points for Flat Top), and is given by:

ENBW / 12.5

(ENBW / 13.28125 for Flat Top window)

See the “Main Time” on page 434 topic for more on relationships between Res BW and time.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

BW

VSA

<meas>:=VECTor|ADEMod|IPOWer|IDEMod|MOTotalk

[:SENSe]:<meas>:BANDwidth|BWIDth[:RESolution]

<bandwidth>

[:SENSe]:<meas>:BANDwidth|BWIDth[:RESolution]?

VECT:BWID 200 KHZ

VECT:BWID?

Key blanked in any other measurement than Vector or Analog Demod

Changing Main Time or Gate Length changes Res BW.

See Res BW Coupling for other changes that can affect (or be affected by) Res

BW

300 kHz

Saved in instrument state.

–9.9e37

9.9e37

Prior to A.02.00

A.02.00

Res BW Coupling

This property controls how Res BW is affected by other parameters. The three possible settings are:

Span: (default) This setting keeps the ratio of Res BW:Span constant whenever the Span is changed.

However, you can change the Res BW at will, and doing so establishes a new Res BW:Span ratio.

369

Common Measurement Functions 2

Min: This setting is only available when the Freq Points property is manually set, and is disabled

(forceful grey out) when Freq Points is Auto. It maintains the RBW at the minimum possible value given the settings for Freq Points, Span, and Window. Res BW coupling is changed from Min to Span if you manually set Res BW.

Fixed: This setting attempts to keep the Res BW setting fixed as Span, Freq Points, or FFT Window type change. Changing FFT Window will cause Main Time (or Gate) length to change in order to keep the Res BW Fixed. Res BW coupling is forced to Fixed mode any time you turn time the Gate

function on or manually set Main Time length. See “Main Time” on page 434 for details.

If a requested change to Res BW or Time Length (Main or Gate) would cause the Res BW to go outside the minimum or maximum Res BW: Span limits (see the main Res BW section for specifics), the Res

BW is clipped at the appropriate limit. The Time length is then set to according to the limited Res BW.

In Fixed coupling mode, if increasing the Span would cause the new Res BW:Span to drop below the minimum, or if decreasing Span would cause the new Res BW: span to exceed the maximum, the requested Span is accepted and then the Res BW is changed to the limiting value. The associated Time length is updated.

In Fixed or Span coupling, increasing Freq Points does not cause the Main (or Gate) Time Length to increase. It only adds zero padding to the array that is used in the FFT to calculate the Spectrum.

Therefore, it will not affect Res BW. If decreasing Freq Points decreases the maximum time length below the current Main Time, then the Main Time length is clipped to the new limits. If Gating is on, the

Gate Delay is first limited, then the Gate Length. The Res BW is then updated as a result of the Time changes.

In Fixed or Span coupling, changing the Window Type will not affect RBW unless it falls outside the limits calculated using the new window. Then the Res BW is clipped at the appropriate limit. The associated Time length is also updated.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Couplings

Preset

State Saved

Range

BW

VSA

<meas>:=VECTor|ADEMod|IPOWer|IDEMod|MOTotalk

[:SENSe]:<meas>:BANDwidth|BWIDth[:RESolution]:COUPle

SPAN|MIN|FIXed

[:SENSe]:<meas>:BANDwidth|BWIDth[:RESolution]:COUPle?

VECT:BWID:COUP FIX

VECT:BWID:COUP?

Blanked when in any other measurement than Vector or Analog Demod

MIN is not available if Freq Points is set to Auto and trying to set it generates error –221 Settings conflict

See narrative above table and also Res BW section

SPAN

Saved in instrument state.

Span | Min | Fixed

370

Common Measurement Functions 2

Initial S/W Revision

Modified at S/W Revision

Prior to A.02.00

A.02.00

FFT Window

This key allows you to choose the Window function that is applied to the time data prior to the FFT calculation used for Spectrum and PSD displays. Four windows are available.

Window name

Uniform

Hanning

Gaussian

Flat Top

Common usage

Transient or self-windowing signals, signals that are periodic within a time record length.

Frequency resolution

High dynamic range

High amplitude accuracy

Normalized ENBW (Hz-s)

1.0

1.5

2.21536

3.8194

The normalized ENBW is the equivalent noise bandwidth, that is, the width of a rectangular filter that passes the same amount of white noise as the window. It is used to define the resolution bandwidth.

Key Path

Mode

Measurement

Remote Command

Example

Couplings

Preset

State Saved

Range

Readback Text

Initial S/W Revision

Modified at S/W Revision

BW

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

[:SENSe]:<meas>:FFT:WINDow[:TYPE]

UNIForm|HANNing|GAUSsian|FLATtop

[:SENSe]:<meas>:FFT:WINDow[:TYPE]?

VECT:FFT:WIND GAUS

VECT:FFT:WIND?

See Res BW and Res BW Coupling sections

FLAT

Saved in instrument state.

Uniform | Hanning | Gaussian (High Dyn Rng) | Flat Top (High Amptd Accy)

Uniform | Hanning | Gaussian | Flat Top

Prior to A.02.00

A.02.00

FREQ Channel

Frequency parameters for any vector measurement consist of 2 pairs of properties: Center Frequency and

371

Common Measurement Functions 2

Span or Start Frequency and Stop Frequency. These behave much as they do in any other application, but there is the additional constraint that the span is limited to much less than the center frequency range.

If you change center frequency the start and stop frequencies change by the same amount.

If you change span, start frequency and stop frequency are changed by 1/2 the span change.

If you change start frequency, stop frequency remains fixed and span and center frequency are refigured accordingly. Changing stop frequency has similar behavior.

Limits:

If you change the start frequency such that it will equal or exceed the stop frequency, the new start frequency will be accepted if possible and the stop frequency will be set to min span above the start.

Similarly if you attempt to set the stop below the start, the start frequency will move to a min span below the new stop frequency.

If you reduce the start frequency beyond a max span below the stop, the stop frequency will be "dragged along" such that it will be a max span above the new start frequency, and similarly increasing the stop frequency will drag the start frequency along if you attempt to increase the span beyond the maximum.

Stop frequency may be 1/2 span above the maximum center frequency, but frequency-domain traces are blanked above the maximum center frequency.

Start frequency may be 1/2 span below the minimum center frequency, but frequency-domain traces are blanked below the minimum center frequency.

Pressing the Freq hardkey changes the active function to Center Frequency.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Front Panel

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Center Freq

Sets the frequency of the display Center.

Key Path

Mode

Remote Command

Example

Couplings

Preset

FREQ Channel

VSA, LTE, LTETDD, IDEN

[:SENSe]:FREQuency:CENTer <freq>

[:SENSe]:FREQuency:CENTer?

FREQ:CENT 985 MHZ

FREQ:CENT?

Start Freq, Stop Freq, and Span. See “FREQ Channel” on page 371

for more details.

1 GHz

372

Common Measurement Functions 2

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Start Freq

Sets the frequency of the display Start.

Saved in instrument state.

0 Hz

Depends on frequency range option.

Prior to A.02.00

A.02.00

Key Path

Mode

Remote Command

Example

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

FREQ Channel

VSA, IDEN

[:SENSe]:FREQuency:STARt <freq>

[:SENSe]:FREQuency:STARt?

FREQ:STAR 980 MHz

FREQ:STAR?

Stop Freq, Center Freq, and Span. See “FREQ Channel” on page 371 for

more details.

Depends on span option. It is 1/2 max span below 1 GHz

Saved in instrument state.

–9.9e37

9.9e37

Prior to A.02.00

A.02.00

Stop Freq

Sets the frequency of the display Stop.

Key Path

Mode

Remote Command

Example

Couplings

FREQ Channel

VSA, IDEN

[:SENSe]:FREQuency:STOP <freq>

[:SENSe]:FREQuency:STOP?

FREQ:STOP 990 MHz

FREQ:STOP?

Start Freq, Center Freq, and Span. See “FREQ Channel” on page 371 for more

details.

373

Common Measurement Functions 2

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Depends on span option. It is 1/2 max span above 1 GHz

Saved in instrument state.

–9.9e37

9.9e37

Prior to A.02.00

A.02.00

CF Step

This key controls the amount the center frequency changes if it is the active function and the user presses the Up or Down arrow key. Note: the start and stop frequency also change by the amount of the CF Step if the Up/Down arrow keys are used to change them; but the key is mainly used is in connection with stepping the center frequency, so the legacy key name has been retained. The step size in Auto mode is

1/10th the span. It can be set to any value in manual mode.

Key Path

Mode

Remote Command

Example

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

FREQ Channel

VSA, LTE, LTETDD, IDEN

[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>

[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?

[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1

[:SENSe]:FREQuency:CENTer:STEP:AUTO?

FREQ:CENT:STEP 1 MHZ

FREQ:CENT:STEP?

FREQ:CENT:STEP:AUTO ON

FREQ:CENT:STEP:AUTO?

1/10th Span when auto is turned on

Depends on span option; 1/10th default span.

Saved in instrument state.

–9.9e37

9.9e37

Prior to A.02.00

A.02.00

Marker

The Marker hardkey displays the Marker menu. A marker can be placed on a trace to allow the value of the trace data at the marker position to be determined precisely. Markers can also be used in pairs to read the difference (or delta) between two data points. They can also be used to make power calculations over

374

Common Measurement Functions 2

a band of frequencies or a time interval. See Marker Functions below for more details.

The functions in this menu include a 1-of-N selection of the control mode Normal, Delta, Fixed, or Off for the selected marker. The control mode is described below.

Pressing Marker always makes the selected maker's X position the active function.

If the currently selected marker is Off, pressing Marker sets it to Normal mode and places it at the center of the screen on the currently selected trace.

As a convenience, if there are no markers displayed on the current trace, pressing the marker hardkey

(whenever the marker menu is already showing) selects the lowest numbered marker that is currently off and turns it on in normal mode on the selected trace. In other words, pressing the Marker hardkey twice will always turn on a marker on the selected trace if none was turned on before.

For more information see the Analyzer Setup, Marker for a description of this function.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Front Panel

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Select Marker

Specifies the selected marker. The selected marker is the one that is affected by the marker position and properties settings, peak search, and other marker functions. Several menus have a Select Marker key for convenience. Marker selection using any one of these is reflected in all others, i.e., there is only one selected marker for the whole measurement. If all markers are off, then marker 1 becomes the selected marker.

As a convenience, if no markers are displayed on the selected trace, selecting a marker that is off automatically turns it on in normal mode on the selected trace.

There is no SCPI function for selecting a marker. Instead, SCPI functions may explicitly include the index of the marker for which they are to apply. (Most SCPI marker functions that affect the state of a marker will also make it the selected marker for front panel commands.)

Key Path

Mode

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Marker or Marker> or Marker Function

VSA, LTE, LTETDD, IDEN

No

1|2|3|4|5|6|7|8|9|10|11|12

Prior to A.02.00

A.02.00

Control Mode

The control mode of the selected marker is selected by pressing

Normal

,

Delta

,

Fixed

, or

Off

. The behavior of a marker under each control mode is described below the table. The current control mode is

375

Common Measurement Functions 2

shown by highlighting the appropriate key.

The SCPI command in the table below selects the marker and sets the marker control mode as described under Normal, Delta, Fixed and Off, below. All interactions and dependencies detailed under the key description are enforced when the remote command is sent.

Key Path

Mode

Measurement

Remote Command

Example

Couplings

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Marker

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MO

DE POSition|DELTa|FIXed|=OFF

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MO

DE?

CALC:VECT:MARK1:MODE POS

CALC:VECT:MARK1:MODE?

When Delta mode is selected, or when the mode is changed from Delta to Off, the marker relative to the selected marker may be affected, as described in the text descriptions below.

=OFF

Saved in instrument state.

Normal|Delta|Fixed|Off

Prior to A.02.00

A.02.00

Normal (Position)

A marker in normal mode reports the trace data value (Y value) at a particular point on a trace. The marker's absolute X (and Z) position is specified by you in displayed units. The marker symbol appears on the trace at the specified position and tracks the absolute Y value at that position as it changes from scan to scan. The absolute Y value is displayed in the marker readout area. In older instruments this was called Position mode, and the designation may still be used for backward compatibility.

For Control Mode SCPI command information see:

“Control Mode” on page 375

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Delta

A marker set to delta mode reports the difference between Y values at two points. A delta marker is

376

Common Measurement Functions 2

relative to an associated reference marker on the same trace. (The reference marker may be set on the

Marker, Properties, Relative To menu). The reference marker is usually fixed, but may also be normal or delta. The X (and Z) position of a delta marker is specified as an offset from the reference marker position. The delta marker symbol tracks the absolute Y value just like a normal marker, but the marker readout displays the difference between the absolute Y values of the delta marker and its reference marker (absolute units are used even if the reference is itself a delta marker). Usually this is a straight difference in the current displayed units. For example, if the trace format is LogMag (dBm), the delta marker displays the difference in dB, thus showing a power ratio. But if the trace format is Real, then the delta marker shows a voltage difference, not a ratio. Exceptions for this are:

• When the trace format is Linear Mag or Log Mag (linear unit) the delta marker displays a voltage ratio, or (if the Y Axis unit is Power) a power ratio, rather than a difference.

• When either the marker or its reference has a marker function turned on, the delta marker always displays a ratio or its decibel equivalent. See Marker Function for more details on how delta markers work with marker functions. The type of ratio calculated (power or voltage) depends on the delta marker units; the reference marker value is converted as needed so it has compatible units.

• When the trace format is Wrap Phase, the delta marker readout is constrained to the wrapped phase display range, which is usually (–180, +180] degrees. For example, if the absolute phase at marker 1 is 170 deg and its reference has phase of –170 deg, the delta will not show 340 deg, but –20 deg. Note that the Wrap Phase display range can be changed (see Trace/Detector, Phase/Delay Properties,

Phase/Trellis Offset).

There is no current support for calculating deltas across traces (and this cannot be done at all unless the traces have the same domain and ranges).

By default, the reference marker for marker 1 is marker 2; for marker 2 is 3 and so on, but the reference marker may be changed. See the section on the "Relative To" softkey below.

The following coupling rules apply from the front panel, and also if the equivalent SCPI commands are sent.

Pressing the Delta key causes the selected marker becomes a delta marker if it is not already. Also, the selected marker's reference is affected as follows:

• If the reference marker was off, it is turned on as a fixed marker.

• The reference marker is moved to the trace of the selected marker and set to the same position as the selected marker.

• If the delta marker has a marker function turned on, the reference marker takes on the same function

(with the same band limits).

Exception: Pressing Delta when the selected marker's mode is not yet Delta does not move or change a reference marker that is already turned on (Normal, Delta, or Fixed) and on the same trace as the selected marker. It merely changes the selected marker's mode to Delta and shows the current offset between it and the reference. If you press Delta again (when the selected marker is already in Delta mode) then the reference is moved and modified as described above.

When a delta marker is changed to any other control mode, if its reference marker is fixed then the reference marker is also turned off.

If you move a delta marker to a different trace, it is forced to Normal mode, and if its reference is fixed, the reference is turned off.

377

Common Measurement Functions 2

A delta marker is forced to Normal mode if you turn its reference off, or if you move its reference to another trace. (In the latter case the reference is not turned off even if it is fixed.)

If you change the selected marker's reference (using the Marker, Properties, Relative To) the selected marker is forced to Delta mode. This change of the selected marker to Delta mode causes its new reference's control mode and position to change as described above.

For Control Mode SCPI command information see:

“Control Mode” on page 375

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Fixed

Fixed markers are mainly used as reference markers for Delta markers. A fixed marker's X and Y Axis values may be directly or indirectly specified by you, and they remain fixed once specified, i.e. they do not follow the trace data value. These markers are represented on the display by an “X” rather then a diamond. If a marker is changed from off to fixed, the X and Y (and Z) values are chosen to put it in the center of the display. If the marker is changed from some other type to fixed, the current X and Z values of the marker remain unchanged. The Y value is taken from the current trace data value and must be changed manually thereafter.

For Control Mode SCPI command information see:

“Control Mode” on page 375

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Off

Turning a marker off makes it invisible, and also its annotation.

Turning a marker on (i.e. changing its control mode from Off to any other control mode) assigns the marker to the currently selected trace.

For Control Mode SCPI command information see:

“Control Mode” on page 375

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

378

Common Measurement Functions 2

Marker Position

Marker position is used to select which data point in a trace we want to read out with the marker (or where to locate a fixed marker). The marker position is primarily set in terms of the domain units, not trace points (although it can be set in terms of points via SCPI). The default active function when you press a marker hard key is the X position for the currently selected marker. The exception to this is when the selected marker is fixed. In that case there is no default active function (to prevent inadvertently changing a fixed marker's location).

Marker position is not defined when a marker's control mode is Off. When a marker is turned on in

Normal or Delta mode, its X (and Z) values are set to the center of the trace data. If a marker is turned on in Fixed mode, its position is set so that it appears in the middle of the trace grid.

The Marker Position key branches to the Marker Position menu, which allows you to set any position variable relevant to the selected marker's control mode and trace format.

For Normal and Delta markers, usually only Marker X is available. Marker Z is available for trace data with 2-dimensional domain. For Fixed markers, Y may also be set. If the trace format is Vector or

Constellation, Marker Y controls the real (horizontal axis) value and Marker Y Imag controls the imaginary (vertical axis) value. The key (or the keys below it) is grayed out if the selected marker is off.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Marker X

This sets the selected marker's X Axis value position in the current X Axis Scale unit. If the control mode is Off, the SCPI command has no effect other than to cause the marker to become selected

.

Note that the X value may change if the marker is moved to a trace with a different domain.

The Marker X position is absolute if the marker control mode is Normal or Fixed. If the control mode is

Delta, then the X position is relative to the reference marker. The valid X positions are the actual data points in the trace; the marker cannot be located between points. If a SCPI command attempts to place the marker between two points, the X value snaps to the closest point.

Note that for Vector or Constellation format, the X axis is perpendicular to the screen (because the screen axes are used to show the real and imaginary parts of the Y value), so adjusting the X value in this case will only cause the marker to move horizontally if the real Y value changes. For Fixed markers on a trace with one of these formats, adjusting the X value will not cause horizontal motion of the marker at all.

Instead, you use the Marker Y and Marker Y (imag) controls to move the marker horizontally and vertically.

Key Path

Mode

Measurement

Marker, Marker Position

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

379

Common Measurement Functions 2

Remote Command

Example

Notes

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X

<real>

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X?

CALC:VECT:MARK:X 0.325

CALC:VECT:MARK:X?

Marker X will not go outside the bounds of the data unless it is Fixed. If you attempt to set it to a value outside the bounds it will be clipped at the closest limit, and error –222 Data Out of Range is generated.

If suffix is sent, it must match the X units for the trace the marker is on.

Otherwise, error –138, "Suffix not allowed" is generated.

If you try to read or set the position of a Delta marker, remember that the position is in relative units.

See Coupling at the end of the Control Mode section. See also Couple

Markers section.

None until marker is turned on.

Saved in instrument state.

Depends on trace data

Depends on trace data

Prior to A.02.00

A.02.00

SCPI only X position commands

Via SCPI , the marker position may also be set or queried in trace points. In this case, the position setting or reading is absolute regardless of control mode.

NOTE

The entered value in Trace Points is immediately translated into the current domain units for setting the value of the marker. The marker's value in domain units, NOT trace points, will be preserved if a change is made to the X Axis scale settings. Thus, if you use this command to place a marker on point 500, which happens at that time to correspond to 13 GHz, and then you change the Start

Frequency so that point 500 is no longer 13 GHz, the marker will stay at 13 GHz,

NOT at point 500.

If the trace the marker is on has a 2-dimensional domain, then the points are numbered in the following way:

Starting at the minimum X and Z position, this point is numbered 0. Each time you increment the point number, increment the X value to the next available value. When X reaches the maximum X position, then reset X to the minimum and increment the Z value. Then continue incrementing the X position in the same manner as before.

Note that for symbol tables, which have no axes, incrementing the X position in points moves the marker

380

Common Measurement Functions 2

consecutively through all table entries.

Mode

Measurement

Remote Command

Example

Notes

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:X

]:POSition <real>

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12[:X

]:POSition?

CALC:VECT:MARK:POS 25

CALC:VECT:MARK:POS?

When a marker control mode is changed from off to any other mode, the X position is set to mid-screen.

Same couplings as for Marker X value

None until marker is turned on.

Saved in instrument state.

Depends on trace data

Depends on trace data

Prior to A.02.00

A.02.00

Marker X Unit may be queried via SCPI.

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X:

UNIT?

CALC:VECT:MARK:X:UNIT?

Query Only

Prior to A.02.00

A.02.00

Marker Y

This function only affects fixed markers. It allows you to set or read back the selected marker's Y Axis value in the current Y Axis Scale unit. Setting the Y value has no effect (other than to cause the marker to become selected) if the control mode is other than fixed. The query form generates an error if the control mode is Off. Note that the Y value may change if the Y-axis units change, either from a change in

381

Common Measurement Functions 2

format of the trace the marker is on, or if the marker is moved to a different trace.

If the selected marker is on a trace that is displayed with Vector or Constellation format, this function controls only the real part of the Y value (i.e., the horizontal axis value). Use the Marker Y (imag) control to change the imaginary (vertical) value. Marker Y and Marker Y Imag always set or get the rectangular form of Y, regardless of whether the marker readout is polar or rectangular.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Marker, Marker Position

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Y[

:REAL] <real>

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Y[

:REAL]?

CALC:VECT:MARK2:Y 0.325

CALC:VECT:MARK2:Y?

You cannot set Y unless the marker type is fixed. If the marker becomes fixed after a marker function is turned on, it is set to whatever the Y value was when the marker became fixed.

If suffix is sent, it must match the Y units for the trace the marker is on.

Otherwise, error –138, "Suffix not allowed" is generated.

Changes if marker is relative to a Delta marker that is turned on or re-zeroed

(see Coupling of Delta and Reference Markers).

None until marker is turned on.

Saved in instrument state.

–9.9E+37

9.9E+37

Prior to A.02.00

A.02.00

Marker Y Unit may be queried via SCPI.

Mode

Measurement

Remote Command

Example

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Y:

UNIT?

CALC:VECT:MARK:Y:UNIT?

382

Common Measurement Functions 2

Notes

Initial S/W Revision

Modified at S/W Revision

Query Only

Prior to A.02.00

A.02.00

Marker Y Imag (Imaginary)

This only affects fixed markers and is only available when the trace format is Vector or Constellation. It allows you to set or read back the selected marker's quadrature (imaginary) Y value in the current Y Axis

Scale unit. It has no effect (other than to cause the marker to become selected) if the control mode is other than fixed, or if the current trace format is not complex. The query form generates an error if it is used for a marker that is not on a complex trace. Marker Y Imag is not affected by whether the marker readout is polar or rectangular.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Marker, Marker Position

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Y:

IMAGinary <real>

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Y:

IMAGinary?

CALC:DDEM:MARK1:Y:IMAG 0.435

CALC:DDEM:MARK1:Y:IMAG?

Grayed out unless the marker is fixed and on a vector display.

If suffix is sent, it must match the Y units for the trace the marker is on.

Otherwise, an Invalid Suffix error is generated. Otherwise, error –138, "Suffix not allowed" is generated. If query is sent while the marker is on a trace whose format is not vector or constellation, NaN (9.91E+37) is returned.

None until marker is turned on.

Saved in instrument state.

Depends on trace format

Depends on trace format

Prior to A.02.00

A.02.00

Marker Z

For markers on traces with a 2-dimensional domain, this sets the selected markers Z Axis value in the current Z Axis Scale unit. In each case the marker that is addressed becomes the selected marker. It has no effect (other than to cause the marker to become selected) if the control mode is

Off

, or if the trace has no Z domain.

Note that the Z value may change or become irrelevant if the marker is moved to a trace

383

Common Measurement Functions 2

with a different Z domain, or no Z domain.

Note that this Z value is affected if the SCPI command to set marker point position is used.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Marker, Marker Position

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Z

<real>

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Z?

CALC:OFDM:MARK:Z 12

CALC:OFDM:MARK:Z?

Marker Z will not go outside the bounds of the data unless it is Fixed. If you attempt to set it to a value outside the bounds it will be clipped at the closest limit, and error –222 Data Out of Range is generated.

If suffix is sent, it must match the Z units for the trace the marker is on.

Otherwise, error –138, "Suffix not allowed" is generated.

See Coupling at the end of the Control Mode section. See also Couple

Markers section.

None until marker is turned on.

Saved in instrument state.

Depends on trace data

Depends on trace data

Prior to A.02.00

A.02.00

Marker Z Unit may be queried via SCPI.

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:Z:

UNIT?

CALC:OFDM:MARK:Z:UNIT?

Query Only

Prior to A.02.00

A.02.00

384

Common Measurement Functions 2

Marker Properties

The Marker Properties key accesses a menu of common marker properties.

Key Path

Mode

Initial S/W Revision

Marker

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

Relative To

This key allows you to specify which marker is used as a reference for the selected marker when the selected marker’s control mode is set to Delta. By default, the reference marker is numerically one higher than the selected marker, that is, marker 1 is relative to marker 2, marker 2 to marker 3, and so on.

Marker 12 by default is relative to marker 1. This key allows you to change the reference marker from the default. Note that a marker cannot be made relative to itself.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Couplings

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Marker, Properties

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:RE

Ference <integer>

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:RE

Ference?

CALC:VECT:MARK2:REF 4

CALC:VECT:MARK2:REF?

The reference marker cannot be the same value as the selected marker, i.e.,. a marker cannot be relative to itself. The currently selected marker will not be an available choice in the relative to selection (i.e. the selected marker will appear grayed out).

When queried a single value will be returned (the specified marker numbers relative marker).

See Coupling of Delta and Reference Markers above. The old reference remains as it was.

2|3|4|5|6|7|8|9|10|11|12|1

Saved in instrument state.

1|2|3|4|5|6|7|8|9|10|11|12

Prior to A.02.00

A.02.00

385

Common Measurement Functions 2

Complex Format

This determines the format for the readout when a marker is placed on a complex display (vector or constellation). The choices are to read out in rectangular or polar coordinates. The readout format applies to the marker display and marker table only; there is no SCPI for reading out the marker value in polar form.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Marker, Properties

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:CF

ORmat RECTangular|POLar

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:CF

ORmat?

CALC:VECT:MARK1:CFOR RECT

CALC:VECT:MARK1:CFOR?

RECT

Saved in instrument state.

Rect|Polar

Prior to A.02.00

A.02.00

Marker Trace

This key allows you to determine the trace to which a marker is assigned. By default, when a marker is turned on it is assigned to the currently selected trace. You may change that assignment using this control.

Key Path

Mode

Measurement

Remote Command

Example

Couplings

Marker, Properties

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:TR

ACe <integer>

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:TR

ACe?

CALC:VECT:MARK3:TRAC 2

CALC:VECT:MARK3:TRAC?

See Couplings of Delta and Reference Markers above..

386

Common Measurement Functions 2

Preset

State Saved

Range

Min

Max

Initial S/W Revision

Modified at S/W Revision

Marker is assigned to currently selected trace when turned on.

Saved in instrument state.

Trace 1|Trace2|Trace 3|Trace 4

1

4

Prior to A.02.00

A.02.00

Marker Count

This key enables the frequency counter algorithm on the selected marker. This algorithm can more precisely determine the frequency of a peak. The marker must be on a frequency domain trace, with data coming from hardware. Place the marker on a peak and enable the frequency counter. The marker readout then shows the calculated frequency rather than the marker X position. Only one marker can be counted at any time. Turning on marker count for any marker turns it off for all other markers.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Marker, Properties

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FC

Ount[:STATe] OFF|ON|0|1

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FC

Ount[:STATe]?

CALC:VECT:MARK:FCO ON

CALC:VECT:MARK:FCO?

Marker must be on a frequency-domain trace and data must be live, not recorded or simulated.

OFF

Saved in instrument state.

Off|On

Prior to A.02.00

A.02.00

The frequency counter result must be read back with the following SCPI command. The Marker X query command will only get the marker's data point position, which will not be as accurate as the frequency counter result.

Key Path

SCPI only

387

Common Measurement Functions 2

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FC

Ount:X?

CALC:VECT:MARK:FCO:X?

Query only. If the marker counter result is unavailable, NaN is returned.

Prior to A.02.00

A.02.00

Marker Table

When the Marker Table is turned on, the display is split into a measurement window and a marker data display window. For each marker which is on, information is displayed in the data display window, which includes the marker number, control mode, trace number, X axis scale, X axis value, and the

Y-axis result. Additional information is shown for markers which have marker functions turned on.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Marker

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer:TABLe[:STATe] OFF|ON|0|1

:CALCulate:<meas>:MARKer:TABLe[:STATe]?

CALC:VECT:MARK:TABL ON

CALC:VECT:MARK:TABL?

OFF

No

Off|On

Prior to A.02.00

A.02.00

Couple Markers

Marker Coupling affects all currently displayed markers. In general when coupling is turned on then all

Normal or Delta markers with the same (or equivalent) domain as the selected marker move in the same manner as the selected marker. Coupling is relative between markers on the same trace (so that their relative positions in the domain are maintained). Coupling can be absolute between markers on different traces that have equivalent domains. That is, they are forced have the same position in the domain, if possible. (As an example of equivalent domains, demodulated symbol positions can be derived from time by using the current symbol rate). When you move the selected marker, then others on related traces

388

Common Measurement Functions 2

track it. This is to allow you to correlate different measurement results. For example, to you can place a marker at a particular symbol time on an error vector magnitude display, and have tracking markers on the symbol table and pre-demod time trace, showing you the symbol value and the actual time-varying signal value at the same point in time.

Absolute coupling is performed only for the lowest numbered Normal or Delta marker on each trace. All other markers on a trace couple relatively. When you turn on marker coupling, the subset of markers that have the same domain as the selected marker track it and all other markers remain at their current location. The absolutely coupled markers within this subset will be moved at this time to match the domain setting of the selected marker, with the relatively coupled markers following accordingly to maintain offsets within their respective traces. Those markers with different domains remain at their current location. When you select a marker with a different domain than the previously selected marker, then the subset of markers with that domain go through the same procedure.

Any marker that coupling would move outside its range of X values, will remain at the closest limiting value until the selected marker moves in such a way as to bring the coupled X value back into range. If the coupled markers are on data that do not have the same domain resolution, then they are positioned as close to each other as possible.

If markers change mode or trace, or trace data is changed below them, the coupling rules are immediately applied to the new set.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Marker

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer:COUPle[:STATe] OFF|ON|0|1

:CALCulate:<meas>:MARKer:COUPle[:STATe]?

CALC:VECT:MARK:COUP ON

CALC:VECT:MARK:COUP?

OFF

Saved in instrument state.

Off|On

Prior to A.02.00

A.02.00

All Markers Off

This function turns all markers off and sets the selected marker to 1.

Key Path

Mode

Marker

VSA, LTE, LTETDD, IDEN

389

Common Measurement Functions 2

Measurement

Remote Command

Example

Initial S/W Revision

Modified at S/W Revision

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer:AOFF

CALC:VECT:MARK:AOFF:

Prior to A.02.00

A.02.00

Marker -> (Marker To)

The Marker -> hardkey provides access to some convenient functions for copying the marker position to a number of frequency and Y-axis scaling parameters. These functions are available from the front panel only. No SCPI is provided, because you can already read the marker position via SCPI and then set any frequency or scaling parameter accordingly, with full accuracy.

Pressing the Marker -> hardkey always makes the selected marker's X position the active function.

If the selected marker is off, pressing the Marker -> hardkey turns on the selected marker in normal mode on the currently selected trace.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Front Panel

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Mkr -> CF (Center Frequency)

This function sets the center frequency equal to the selected marker's absolute frequency. The marker must be on a frequency-domain trace. The absolute marker frequency is used regardless of whether its control mode is Normal, Delta, or Fixed.

If the currently selected marker is not on when this key is pressed, it will be turned on at the center of the screen as a normal type marker.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker To

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Mkr -> CF Step

This function sets the center frequency step size equal to the selected marker's frequency. The marker must be on a frequency-domain trace. The absolute marker frequency is used regardless of whether its control mode is Normal, Delta, or Fixed.

If the currently selected marker is not on when this key is pressed, it will be turned on at the center of the

390

Common Measurement Functions 2

screen as a normal type marker.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker To

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Mkr -> Start

This function sets the start frequency equal to the selected marker's frequency. The marker must be on a frequency-domain trace. The absolute marker frequency is used regardless of whether its control mode is

Normal, Delta, or Fixed.

If the currently selected marker is not on when this key is pressed, it will be turned on at the center of the screen as a normal type marker.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker To

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Mkr -> Stop

This function sets the stop frequency equal to the selected marker's frequency. The marker must be on a frequency-domain trace. The absolute marker frequency is used regardless of whether its control mode is

Normal, Delta, or Fixed.

If the currently selected marker is not on when this key is pressed, it will be turned on at the center of the screen as a normal type marker.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker To

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Mkr Delta -> Span

This function sets the start and stop frequencies to equal to the selected marker's frequency and that of its reference. That is, the measurement span is "zoomed in" so that the selected marker and its associated reference appear on the extreme left and right of the display. The marker must be on a frequency-domain trace and its control mode must be Delta.

Key Path

Mode

Marker To

VSA, LTE, LTETDD, IDEN

391

Common Measurement Functions 2

Initial S/W Revision

Modified at S/W Revision

Prior to A.02.00

A.02.00

Mkr -> Ref Lvl

This function sets the Y axis reference value equal to the selected marker's Y value. For example, if the reference position is at the top of the screen, the whole trace is moved up so that the marker appears at the top of the screen. Note that this is a display scaling function only. The input range remains the same.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker To

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Counter -> CF (Center Frequency)

This function copies the frequency of the marker counter to the center frequency. The marker counter function must be on.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker To

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Mkr Delta -> CF (Center Frequency)

This function sets the center frequency equal to the difference in frequency between the selected Delta marker and its reference. The marker must be on a frequency-domain trace and the selected marker's control mode must be Delta.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker To

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Marker Function

This key accesses a menu of selectable marker functions for VSA based measurements.

Marker Functions perform post-processing operations on marker data. Band Functions are Marker

Functions that allow you to define a band of frequencies around the marker. The band defines the region of data used for the numerical calculations. These marker functions also allow you to perform mathematical calculations on trace and marker data and report the results of these calculations in place of the normal marker result.

392

Common Measurement Functions 2

Unlike regular markers, marker function markers are not placed directly on the trace. They are placed at a location which is relative to the result of the function calculation.

The Marker Function menu gives you access to power calculations in bands of frequencies or time intervals centered on a marker. It also allows you to make calculations like carrier to noise by combining delta markers with marker functions. Marker functions are generally available for time and frequency domain traces, and not for others. If the marker function calculation is undefined for particular trace data, then "---" is shown in place of a number in the result display and marker table, and

CALC:<meas>:MARK[n]:Y? will return 9.91E+37 (NaN).

Pressing Marker Function always makes the selected marker's X position the active function.

If the selected marker is off, pressing the Marker Function hardkey turns on the selected marker in normal mode on the currently selected trace.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Marker Function

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FU

NCtion BPOWer|BDENsity|=OFF

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FU

NCtion?

CALC:VECT:MARK1:FUNC BPOW

CALC:VECT:MARK1:FUNC?

:CALC:<meas>:MARK1:FUNC? returns the current function type for marker

1. To return the result, use :CALC:<meas>:MARK1:Y?

=OFF

Saved in instrument state.

Band Power|Band Density|Off

Prior to A.02.00

A.02.00

Band/Interval Power

Turns on the Band/Interval Power function for the selected marker. This function calculates the power within the band centered on the marker. The function works generally with frequency spectra, PSD and time traces. On traces where band power is undefined, the result display shows "---" and

CALC:<meas>:MARK[n]:Y? will return 9.91E+37 (NaN), although the band interval can still be defined.

Frequency-domain data

If the marker is on a frequency-domain trace, the result is total power within the band. This is true whether the underlying trace data is a power spectrum or power spectral density.

393

Common Measurement Functions 2

Time-domain data

If the marker is on a time-domain trace, the result is average power within the time interval, that is, the power at each time sample in the time interval is calculated, the powers are summed and the total divided by the number of samples.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker Function

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Band Power Calculation

A band/interval power calculation result may be shown in dBm, dBVrms, Watts, Volts RMS Squared or

Volts RMS. The table below shows the choice of display units if Band Power Calculation is set to Mean, depending on the current format and Y units of the trace the marker is on.

Trace data type Trace Format Y Unit

Spectrum, PSD, Time record

Dimensionless (e.g.,

Frequency response,

Impulse response, various

Demodulation error types)

General dimensions(e.g.,

Hz, %)

LogMag (dB)

Linear Mag, Real, Imag,

Log Mag (lin)

Linear Mag, Real, Imag,

Log Mag(lin)

Wrap Phase, Unwrap

Phase, Delay

Vector, Constellation, Eye,

Trellis

LogMag (dB)

Linear Mag, Real, Imag,

Wrap Phase, Unwrap

Phase, Delay, Log Mag

(lin)

LogMag (dB)

Linear Mag, Real, Imag,

Wrap Phase, Unwrap

Phase, Delay, Log Mag

(lin)

Auto, Power

Peak, RMS mRMS

Auto, Peak, RMS, mRMS

Power

Any

Any

Any

Any

Any

Any

Result format

dBm dBVrms dBmVrms

Vrms^2

W

Vrms^2 blanked dBrms rms^2 dB<unit>rms

<unit>rms^2

If the Band Power Calculation is set to RMS, then the readout unit does not depend on trace format or Y unit. For Spectrums, PS and Time record traces the displayed unit is "Vrms" For general units, the unit

394

Common Measurement Functions 2

abbreviation is shown followed by "rms".

The Band Power Calculation only controls the readout format for Normal and Fixed markers. For Delta markers, see the discussion below.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Marker Function, Band/Interval Power

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FU

NCtion:BPOWer:CTYPe MEAN|RMS

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FU

NCtion:BPOWer:CTYPe?

CALC:VECT:MARK1:FUNC:BPOW:CTYP MEAN

CALC:VECT:MARK1:FUNC:BPOW:CTYP?

MEAN

Saved in instrument state.

Mean|RMS

Prior to A.02.00

A.02.00

Band/Interval Density

This function calculates the average power density within the band centered on the marker. The function works generally with frequency spectra, PSD and time traces. On traces where band power cannot reasonably be defined, the result display shows "---" and CALC:<meas>:MARK[n]:Y? returns NaN

(9.91E+37), although the band interval can still be defined.

Frequency-domain data

If the marker is on a frequency-domain trace, the result is the band power (as computed above) divided by the bandwidth over which it is measured. This is true whether the underlying trace data is a power spectrum or power spectral density.

Time-domain data

If the marker is on a time-domain trace, the result is average power within the time interval (as computed above) divided by the equivalent noise bandwidth of the span.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker Function

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

395

Common Measurement Functions 2

Band Density Calculation

Turns on the Band/Interval Density function for the selected marker. If the selected marker is off, it is turned on in Normal marker mode and located at the center of the screen.

When Band/Interval Density is selected while in the Marker Function Off state, the Band Span or

Interval Span is initialized to 5% of the screen width.

If the detector mode for the detector on the marker's trace is set to Auto, the average detector is selected.

If the Average type is set to Auto, Power Averaging is selected. Other choices for the detector or Average type will usually cause measurement inaccuracy.

A band/interval density calculation result may be shown in dBm/Hz, Volts RMS Squared or Volts RMS.

The table below shows the choice of display units if Band Density Calculation is set to Mean, depending on the current format of the trace the marker is on.

Trace data type

Spectrum, PSD, Time record

Dimensionless (e.g., Frequency response, Impulse response, various Demodulation error types)

Trace Format Result format

LogMag (dB)

Linear Mag, Real, Imag, Wrap

Phase, Unwrap Phase, Delay, Log

Mag (lin)

LogMag (dB)

Linear Mag, Real, Imag, Wrap

Phase, Unwrap Phase, Delay, Log

Mag (lin) dBm/Hz

Vrms^2/Hz dBrms/Hz rms^2/Hz

General dimensions(e.g., Hz, %) LogMag (dB)

Linear Mag, Real, Imag, Wrap

Phase, Unwrap Phase, Delay, Log

Mag (lin) dB<unit>rms/Hz

<unit>rms^2/Hz

If the Band Density Calculation is set to RMS, then the readout unit does not depend on trace format.

For Spectrum, PSD and Time record traces the displayed unit is "Vrms/rtHz" For general units, the unit abbreviation is shown followed by "rms/rtHz ".

The Band Density Calculation only controls the readout format for Normal and Fixed markers. For Delta markers, see the discussion below.

Key Path

Mode

Measurement

Remote Command

Marker Function, Band/Interval Power

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FU

NCtion:BDENsity:CTYPe MEAN|RMS

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FU

NCtion:BDENsity:CTYPe?

396

Common Measurement Functions 2

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

CALC:VECT:MARK1:FUNC:BDEN:CTYP RMS

CALC:VECT:MARK1:FUNC:BDEN:CTYP?

MEAN

Saved in instrument state.

Mean|RMS

Prior to A.02.00

A.02.00

Band Adjust

These keys allow you to define the bandwidth around the marker. The band is always centered on the marker position. Entering the menu always sets Band/Interval Span as the active function

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker Function

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Band/Interval Center

This function defines the center of the band. That is, it allows you to adjust the marker position in absolute units (regardless of whether the marker mode is Normal or Delta).

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Min

Max

Initial S/W Revision

Marker Function, Band Adjust

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FU

NCtion:BAND:CENTer <real>

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FU

NCtion:BAND:CENTer?

CALC:VECT:MARK2:FUNC:BAND:CENT 1.23E+09

CALC:VECT:MARK2:FUNC:BAND:CENT?

Center of screen

Saved in instrument state.

–9.9E+37

9.9E+37

Prior to A.02.00

397

Common Measurement Functions 2

Modified at S/W Revision A.02.00

Band/Interval Span

Sets the width of the span for the selected marker. This function defines the span of frequencies or time.

The marker position does not change when you adjust the span.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Marker Function, Band Adjust

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FU

NCtion:BAND:SPAN <real>

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FU

NCtion:BAND:SPAN?

CALC:VECT:MARK2:FUNC:BAND:SPAN 1.23E+06

CALC:VECT:MARK2:FUNC:BAND:SPAN?

When marker turned on, 1/20th of current span or displayed time length

Saved in instrument state.

–9.9E+37

9.9E+37

Prior to A.02.00

A.02.00

Band/Interval Left

This function adjusts the left side of the band. In order to remain centered in the band, the marker position must also change as you change the left edge. The right edge is unaffected.

Key Path

Mode

Measurement

Remote Command

Example

Couplings

Marker Function, Band Adjust

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FU

NCtion:BAND:LEFT <real>

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FU

NCtion:BAND:LEFT?

CALC:VECT:MARK2:FUNC:BAND:LEFT 1.23E+06

CALC:VECT:MARK2:FUNC:BAND:LEFT?

Changes marker X to keep the marker centered in the band

398

Common Measurement Functions 2

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

When marker turned on, 1/40th of current span or displayed time length left of the marker position

Saved in instrument state.

–9.9E+37

9.9E+37

Prior to A.02.00

A.02.00

Band/Interval Right

This function adjusts the right side of the band. In order to remain centered in the band, the marker position must also change as you change the right edge. The left edge is unaffected.

Key Path

Mode

Measurement

Remote Command

Example

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Marker Function, Band Adjust

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FU

NCtion:BAND:RIGHt <real>

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FU

NCtion:BAND:RIGHt?

CALC:VECT:MARK2:FUNC:BAND:RIGHT 1.23E+06

CALC:VECT:MARK2:FUNC:BAND:RIGHT?

Changes marker X to keep the marker centered in the band

When marker turned on, 1/40th of current span or displayed time length right of the marker position

Saved in instrument state.

–9.9E+37

9.9E+37

Prior to A.02.00

A.02.00

Band Power and Delta Markers

When either a Delta marker or its reference has a band power function turned on, the Delta marker readout always shows a ratio calculation. This allows you to perform common calculations like carrier to noise ratio or adjacent channel power ratio. The form of the ratio depends on the main marker function calculation type (Mean or RMS). If the main marker function calculation type is Mean, then when you change the marker to Delta the result will be a power ratio. If the main marker function calculation type is RMS, then the Delta marker result will be a voltage ratio. (If the main marker band power function is

399

Common Measurement Functions 2

off, then the form of the ratio depends on the reference marker calculation type: If it is Mean you get a power ratio, and if it is RMS you get a voltage ratio.)

For example, if the main marker function is Band/Interval Power with a calculation type of Mean and the reference marker function is Band/Interval Power with a calculation type of RMS, then the Delta marker will show the ratio of the main marker “Band/Interval Power Mean” value to the reference marker

“Band/Interval Power Mean” (not RMS) value.

A dimensionless ratio (for example Volt/Volt or Watt/Watt) is shown with units of "x". The marker function calculation type will tell you whether the ratio is voltage or power (see above). A dimensionless power ratio is shown with units of dB if the trace format is Log Mag (dB)

If the reference marker function is Band/Interval Density and the main marker is either Band/Interval

Power or its function is turned off, then the ratio is not dimensionless, but has units of Hz (or dB-Hz) for power calculations, or rtHz for voltage calculations. When the main marker function is Band/Interval

Density and the reference is either Band/interval Power or its function is off, the units are /Hz (or dB/Hz) for power calculations, or /rtHz for voltage calculations.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Marker Function

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Meas Setup

This key accesses a menu of keys that select measurement functions for VSA based Measurements.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Front Panel

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Avg Number

This key allows you to turn averaging on or off, and set the number of scans (time records) whose measurement results will be averaged. Averaging can be done over spectrum results (RMS) or over time records (Time). A third kind of pseudo averaging displays the maximum value seen at each spectral line over the specified number of scans. See Average Type for a more detailed description of how measurement results are averaged.) For RMS or Time averaging, the process is similar. Each time an averaged result is displayed, it is the sum of the individual results taken since measurement restart, divided by the number of scans. (For Max averaging, there is no actual summation or division.) The

Measurement Bar shows the number of scans and the Avg number setting; for example, if 4 scans have been taken and the Avg Number is 10, the Meas Bar shows "4/10". The measurement continues to take new scans until the number of scans is equal to the Avg Number setting, at which time the measurement stops if Sweep control is in Single Mode. Otherwise, the measurement continues, and the Average Mode setting determines how successive scans are added to the averaged result. See the Average Mode topic

400

Common Measurement Functions 2

for details.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Meas Setup, More

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

[:SENSe]:<meas>:AVERage:COUNt <integer>

[:SENSe]:<meas>:AVERage:COUNt?

[:SENSe]:<meas>:AVERage[:STATe] OFF|ON|0|1

[:SENSe]:<meas>:AVERage[:STATe]?

VECT:AVER:COUN 20

VECT:AVER:COUN?

VECT:AVER ON

VECT:AVER?

If an averaged measurement is idle because the scan count is equal to the Avg

Number, and the Avg Number is increased, the measurement will resume until the new number of averages is satisfied.

10

OFF

IPOW: ON

Saved in instrument state.

1

2147483647

Prior to A.02.00

A.02.00

Average Mode

The Average Mode determines what happens when the Sweep Mode is Continuous and the number of scans processed exceeds the Average Number (see above). If the Sweep Control is in Single mode, this setting has no effect.

When averaging is on and the number of scans is less than or equal to the Avg Number setting, a linear average is calculated as explained in the Avg Number topic. After the scan count exceeds the Avg

Number setting, the measurement continues to take new scans. The Measurement Bar average indicator shows ">N/N" where N is the Avg Number.

If Average Mode is Exp then new results are averaged in exponentially. In other words, each succeeding average will be the weighted sum of the previous average, weighted by (N–1)/N, and the new measurement, weighted by 1/N, where N is the Average Number setting. (For Max averaging, no weighting occurs; the result continues to be the max value seen at each spectral line for every previous

401

Common Measurement Functions 2

scan since measurement restart.)

If Average Mode is Repeat, then the average buffer will be cleared after the average counter reaches the

Average Number setting, and the average counter will be reset to 0. Then a new set of averages is taken.

The measurement bar therefore continues to show "k/N" in the average indicator, where k is the number of scans since the last time the average buffer was cleared and N is the Avg Number. The averaged result is the sum of the last k results divided by k. (For Max averaging, no sum or division takes place, but the buffer is cleared as stated above. The averaged result is the max value seen over the last k scans.)

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Meas Setup

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

[:SENSe]:<meas>:AVERage:TCONtrol EXPonential|REPeat

[:SENSe]:<meas>:AVERage:TCONtrol?

VECT:AVER:TCON EXP

VECT:AVER:TCON?

EXP

Saved in instrument state.

Exp|Repeat

Prior to A.02.00

A.02.00

Average Setup

This key accesses a menu allowing you to set Averaging parameters for all VSA based measurements.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Meas Setup

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Average Type

The Average Type key allows you to select the type of averaging. The table below shows what measurement results are averaged for each average type. This applies in the Vector Measurement. For the

402

Common Measurement Functions 2

other measurements, details of additional results that is averaged will be given in the measurement PD.

Average Type

RMS

Time

Max

Measurement result averaged

Spectrum, PSD: Power is averaged for each spectral line (i.e., this is a mean-square average of voltage). For the Spectrum result only, if the display transform is linear or real, the RMS result is displayed.

Main Time: Individual time samples in the current time record are averaged vectorially (not RMS) with corresponding points in previous time records. See

“Main Time” on page 434 for more details.

Spectrum, PSD: Not strictly an average. For each spectral line, power from the current measurement is compared to the average buffer value and the maximum is kept in the average buffer.

Some measurement results are inherently averaged, and are not affected by the Average controls. These are: CCDF, CDF, and PDF. They average continuously until the next measurement restart.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Meas Setup, Average Setup

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

[:SENSe]:<meas>:AVERage:TYPE RMS|TIME|MAXimum

[:SENSe]:<meas>:AVERage:TYPE?

VECT:AVER:TYPE RMS

VECT:AVER:TYPE?

RMS

Saved in instrument state.

RMS|Time|Max

Prior to A.02.00

A.02.00

Fast Average

Fast average controls the display of average data. If fast averaging is off, then the display is updated after each time record is processed. If fast averaging is on, then the display is only updated after every M records, where M is the Update Rate (see below). For example, if the fast average count is 10, then the running average is only displayed every 10th time record.

Key Path

Mode

Meas Setup, Average Setup

VSA, LTE, LTETDD, IDEN

403

Common Measurement Functions 2

Measurement

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

[:SENSe]:<meas>:AVERage:FAST OFF|ON|0|1

[:SENSe]:<meas>:AVERage:FAST?

VECT:AVER:FAST ON

VECT:AVER:FAST?

OFF

Saved in instrument state.

On|Off

Prior to A.02.00

A.02.00

Update Rate

The Update Rate controls how often the display updates when fast averaging is turned on. If the Fast

Averaging State is MAX then the display is updated only after the full Average Count is reached.

Otherwise, the display is updated whenever the average count is a multiple of the Update Rate..

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Min

Max

Initial S/W Revision

Meas Setup, More, Average Setup

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

[:SENSe]:<meas>:AVERage:FAST:URATe <integer>

[:SENSe]:<meas>:AVERage:FAST:URATe?

[:SENSe]:<meas>:AVERage:FAST:URATe:AUTO OFF|ON|0|1

[:SENSe]:<meas>:AVERage:FAST:URATe:AUTO?

VECT:AVER:FAST:URAT 20

VECT:AVER:FAST:URAT?

VECT:AVER:FAST:URAT:AUTO ON

VECT:AVER:FAST:URAT:AUTO?

10

ON

Saved in instrument state.

1

2147483647

Prior to A.02.00

404

Common Measurement Functions 2

Modified at S/W Revision A.02.00

PhNoise Opt

Allows you to adjust the LO phase noise optimization to give better close-in phase noise, or better wide-offset phase noise. The definition of what frequency offsets constitute close in or wide offset varies with hardware. (The selection keys provide hardware-specific prompts.)

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Notes

Meas Setup

VSA, WIMAXFIXED

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM

[:SENSe]:<meas>:FREQuency:SYNThesis[:STATe] 1|2

[:SENSe]:<meas>:FREQuency:SYNThesis[:STATe]?

VECT:FREQ:SYNT 1

VECT:FREQ:SYNT?

all VXAmeasurements: Best Wide-offset

Φ Noise

WIMAXFIXED EVM measurement: Best Close-in

Φ Noise

Yes

Parameter key:

1 - optimizes phase noise for close-in frequencies

2 - optimizes phase noise for wide-offset frequencies

The softkey shows the options more explicitly. For MXA/EXA, the selection keys show the options:

Best Close-in

Φ Noise [offset < 20 kHz]

Best Wide-offset

Φ Noise [offset > 30 kHz]

For PXA the options are:

Best Close-in

Φ Noise [offset < 180 kHz]

Best Wide-offset

Φ Noise [offset > 210 kHz]

A.04.00

Initial S/W Revision

Best Close-in

Φ Noise

The LO phase noise is optimized for smaller offsets from the carrier, at the expense of phase noise farther out. For MXA, close in means offsets < 20 kHz; for PXA, it means offsets < 140 kHz.

Key Path

Mode

Meas Setup, PhNoise Opt

VSA, WIMAXFIXED

405

Common Measurement Functions 2

Example VECT:FREQ:SYNT 1

VECT:FREQ:SYNT?

A.04.00

Modified at S/W Revision

Best Wide-offset

Φ Noise

The LO phase noise is optimized for wider offsets from the carrier, at the expense of phase noise closer in. For MXA, wide-offset means offsets > 30 kHz; for PXA, it means offsets > 160 kHz.

Key Path

Mode

Example

Modified at S/W Revision

Meas Setup, PhNoise Opt

VSA, WIMAXFIXED

VECT:FREQ:SYNT 2

VECT:FREQ:SYNT?

A.04.00

Peak Search

The Peak Search hardkey displays a menu that allows markers to be easily moved among peaks on a trace, and also performs the peak search function, as described below. Pressing Peak Search also makes the selected marker's X position the active function.

The peak search function causes the marker to move to the highest point in the trace. The highest point is the point with the largest y-axis value in the current trace format. If the format is complex (vector or constellation) then the point with the highest magnitude is chosen.

Pressing the Peak Search hard key always performs a Peak Search, with one exception: if the Peak

Search menu is not showing but the selected marker is on (Normal, Delta, or Fixed), then pressing the

Peak Search hardkey only displays the Peak Search menu. This allows you to select one of the other peak search functions without disturbing the selected marker's position. If you want to perform a peak search in this case, press the Peak Search hardkey again.

If the selected marker is Off, then pressing the Peak Search hardkey once not only shows the menu, but it turns on the selected marker in Normal mode, assigns it to the selected trace, and performs a peak search.

If any peak search SCPI command is invoked on a marker that is Off, the marker is first turned on in

Normal mode and assigned to the selected trace. Then the peak search is performed.

Key Path

Mode

Measurement

Remote Command

Example

Peak Search (press hardkey twice if menu is not showing)

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MA

Ximum

CALC:VECT:MARK2:MAX

406

Common Measurement Functions 2

Notes There is no softkey for this function. Instead, you press the Peak Search hardkey twice. (Pressing it once is sufficient if the Peak Search menu is showing, but twice guarantees that the function will be invoked)

If peak search function is not invoked (because the response to pressing the hardkey was only to show the menu) then the following message is shown:

"Press Peak Search again to perform a Peak Search."

Prior to A.02.00

A.02.00

Initial S/W Revision

Modified at S/W Revision

Next Peak (Next Lower Amptd)

This command moves the marker to the peak next lower in Y value than the peak it is currently on. If the format is complex (vector or constellation) then the marker moves to the closest point that has a lower magnitude than the marker's current position. If this function is invoked via SCPI on a marker that is off, the result is the same as if you sent a Peak Search command.

Key Path

Mode

Measurement

Remote Command

Example

Initial S/W Revision

Modified at S/W Revision

Peak Search

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MA

Ximum:NEXT

CALC:VECT:MARK2:MAX:NEXT

Prior to A.02.00

A.02.00

Next Higher Amptd

This command moves the marker to the peak next higher in Y value than the peak it is currently on. . If the format is complex (vector or constellation) then the marker moves to the closest point that has a higher magnitude than the marker's current position. If this function is invoked via SCPI on a marker that is off, the result is the same as if you sent a Peak Search command.

Key Path

Mode

Measurement

Remote Command

Example

Initial S/W Revision

Peak Search

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MA

Ximum:PREVious

CALC:VECT:MARK2:MAX:PREV

Prior to A.02.00

407

Common Measurement Functions 2

Modified at S/W Revision A.02.00

Next Right

This command moves the marker to the next peak to the right of its current position. If the format is complex (vector or constellation) then the marker moves forward in time to the next peak. If this function is invoked via SCPI on a marker that is off, the result is the same as if you sent a Peak Search command.

A valid peak is one for which the displayed Y-axis values drop monotonically on both sides of the local maximum at least 4% of the distance between the top and bottom of the display grid before the values begin to rise again.

Key Path

Mode

Measurement

Remote Command

Example

Initial S/W Revision

Modified at S/W Revision

Peak Search

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MA

Ximum:RIGHt

CALC:VECT:MARK2:MAX:RIGH

Prior to A.02.00

A.02.00

Next Left

This command moves the marker to the next peak to the left of its current position. If the format is complex (vector or constellation) then the marker moves back in time to the next peak. If this function is invoked via SCPI on a marker that is off, the result is the same as if you sent a Peak Search command.

A valid peak is one for which the displayed Y-axis values drop monotonically on both sides of the local maximum at least 4% of the distance between the top and bottom of the display grid before the values begin to rise again.

Key Path

Mode

Measurement

Remote Command

Example

Initial S/W Revision

Modified at S/W Revision

Peak Search

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MA

Ximum:LEFT

CALC:VECT:MARK2:MAX:LEFT

Prior to A.02.00

A.02.00

408

Common Measurement Functions 2

Mkr -> CF (Center Frequency)

This key is a duplicate of the key of the same name in the Mkr -> menu. It is placed in this menu as a convenience. See the description in the Marker To section.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Peak Search

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Continuous Peak Search

This key turns on Continuous Peak Search for the selected marker. This function be turned on for any marker independently of any other marker. This function moves the marker to the highest point on the trace each time the trace is updated. If the SCPI command refers to a marker that is off, it is turned on in

Normal mode.

It is possible to have Couple Markers and Continuous Peak Search both on. If this is the case, it is recommended that Continuous Peak search be turned on for only one marker in any tracking set (that is, any set of markers with the same or equivalent domain). Otherwise, conflicts over marker position may arise that cause erratic marker movement.

Key Path

Mode

Measurement

Remote Command

Example

Couplings

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Peak Search

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:CP

Search[:STATe] ON|OFF|1|0

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:CP

Search[:STATe]?

CALC:VECT:MARK1:CPS ON

The Continuous Peak Search key is grayed out when the selected marker is a

Fixed

marker. Also, if Continuous Peak Search is on and the selected marker becomes a fixed marker, then Continuous Peak Search is turned off and the key grayed out.

Continuous Peak Search is turned off when the selected marker is turned off.

OFF

Saved in instrument state.

Off|On

Prior to A.02.00

A.02.00

409

Common Measurement Functions 2

Min Search

This command moves the marker to the lowest Y value on the trace. If the format is complex (vector or constellation) then the marker moves to the lowest value in magnitude. If the SCPI command refers to a marker that is off, it is first turned on in Normal mode and then set on the minimum point.

Key Path

Mode

Measurement

Remote Command

Example

Initial S/W Revision

Modified at S/W Revision

Peak Search

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:MI

Nimum

CALC:VECT:MARK2:MIN

Prior to A.02.00

A.02.00

Mkr -> Ref Lvl (Reference Level)

This function sets the Y axis reference value equal to the selected marker's Y value. For example, if the reference position is at the top of the screen, the whole trace is moved up so that the marker appears at the top of the screen. Note that this is a display scaling function only. The input range remains the same.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Peak Search

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Recall

Operation of this key is identical across several measurements. For more details about this key, see

“Recall” on page 139

.

State

Key Path

Mode

Data (Import)

Recall

VSA, LTE, LTETDD, IDEN

Key Path

410

Recall

Common Measurement Functions 2

Mode VSA, LTE, LTETDD, IDEN

Import Trace Data

This selection allows you to import previously saved trace data into a Data Register and optionally display it. Selecting this key displays a menu that allows you to select the destination data register, and also allows you to choose whether or not to display the recalled data in the currently selected trace. After making these selections depress Open… and use the file dialog to select the file you wish to recall.

Recalling trace data into an already used Data Register overwrites the previous data. If the data register is displayed on any trace, the display is updated to reflect the new data.

The SCPI command

:MMEM:LOAD:TRAC:DATA D1|D2|D3|D4|D5|D6,<filename> recalls data into a specified register, but does not display it in the selected trace. Use the command

:DISP:<meas>:TRAC<n>:FEED D1|D2|D3|D4|D5|D6 to display the register in the desired trace.

It is possible to recall trace data saved by other VXA measurements, or measurements made using the

LTE, LTETDD, iDEN, or 89601 applications.

Key Path

Mode

Remote Command

Example

Notes

State Saved

Recall, Data (Import)

VSA, LTE, LTETDD, IDEN

:MMEMory:LOAD:TRACe:DATA

D1|D2|D3|D4|D5|D6,<filename>[,CSV|TXT|SDF|MAT4|MAT|HDF5

|BIN]

:MMEM:LOAD:TRAC:DATA D1,"Trc1.txt",TXT

The Open: dialog box has the following filter options when you are recalling trace data::

• CSV (Comma delimited) (*.csv)

• SDF (Fast) (*.sdf;*.dat)

• Text (Tab delimited) (*.txt)

The file must have the same format as that created by the Export Recorded

Data command.

The SCPI command has an optional file format parameter. If you do not include this parameter in the SCPI command, the file format is determined by the file name extension. If no file extension is recognized, the file is scanned to determine the format.

If you are not licensed to recall a particular file type, then error –203.9010 will be returned. If the file format cannot be determined or the file cannot be recalled successfully, then error –250.5290 is returned. If the recall is successful, then advisory 0.1600 is shown.

No

411

Common Measurement Functions 2

Readback Data 1|Data 2|Data 3|Data 4|Data 5|Data 6

Data 1

Selects the Data 1 register as the destination for the imported data

Key Path

Mode

Recall, Data (Import), Trace (to)

VSA, LTE, LTETDD, IDEN

Data 2

Selects the Data 2 register as the destination for the imported data

Key Path

Mode

Recall, Data (Import), Trace (to)

VSA, LTE, LTETDD, IDEN

Data 3

Selects the Data 3 register as the destination for the imported data

Key Path

Mode

Recall, Data (Import), Trace (to)

VSA, LTE, LTETDD, IDEN

Data 4

Selects the Data 4 register as the destination for the imported data

Key Path

Mode

Recall, Data (Import), Trace (to)

VSA, LTE, LTETDD, IDEN

Data 5

Selects the Data 5 register as the destination for the imported data

Key Path

Mode

Recall, Data (Import), Trace (to)

VSA, LTE, LTETDD, IDEN

Data 6

Selects the Data 6 register as the destination for the imported data

Key Path

Mode

Recall, Data (Import), Trace (to)

VSA, LTE, LTETDD, IDEN

412

Common Measurement Functions 2

Display in Selected Trace

Allows you to select whether the recalled trace data is to be displayed in the current Trace.

Key Path

Mode

State Saved

Recall, Data (Import), Trace (to)

VSA, LTE, LTETDD, IDEN

No

Open …

Key Path

Mode

Recall, Data (Import)

VSA, LTE, LTETDD, IDEN

Save

See “Save” on page 155 in the section "Common Measurement Functions" for more information.

State

Key Path

Mode

Data (Export)

Save

VSA, LTE, LTETDD, IDEN

Key Path

Mode

Save

VSA, LTE, LTETDD, IDEN

Export Trace Data

This selection allows you to export trace data with (optional) associated headers. Selecting this key displays a menu that allows you to choose which Trace to save (default is the selected Trace) and whether or not to save headers with the data. The header information is used by the VXA application when saved trace data is recalled, and allows it to be displayed with the same formatting and scaling that it had when saved. If headers are not saved, the scaling and format are set to default values when the trace is recalled. After making these selections, press Save As… and use the file dialog to choose a file name and format for the saved data.

Trace data may be exported in several different formats. Text and comma-separated variable (CSV) formats are useful for viewing the data or importing it to a spreadsheet program. The other formats are binary and thus more compact. Trace data files may be recalled for viewing into other VXA, LTE,

413

Common Measurement Functions 2

LTETDD, iDEN, or 89601 measurements.

Key Path

Mode

Remote Command

Example

Notes

State Saved

Readback

Save, Data (Export)

VSA, LTE, LTETDD, IDEN

:MMEMory:STORe:TRACe:DATA

TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6,"<filename>"[

,CSV|TXT|SDF|MAT4|MAT|HDF5|BIN[,OFF|ON|0|1]]

:MMEM:STOR:TRAC:DATA TRACE1,"TRC1.TXT",TXT,ON

The Save As… dialog box has the following format options when you are saving trace data:

• CSV (Comma delimited) (*.csv)

• SDF (Fast) (*.sdf;*.dat)

• Text (Tab delimited) (*.txt)

File format saved depends on selection. The appropriate file extension is appended to the filename if it is not supplied by the user.

If the SCPI command includes just a file name, the file format is determined by the filename extension, which must be one of the choices above. *.sdf, or an unrecognized extension chooses the SDF fast format. If the optional file format enumerator is included in the command, then this determines the file format and the file extension is ignored. The optional binary parameter determines if file headers are saved. The default is ON. If file headers are not wanted use the optional “,OFF” parameter.

The optional Boolean parameter determines whether headers are saved in the file. By default the headers are saved.

If you are not licensed to save a particular file type, then error –203.9010 will be returned.If an invalid file format is specified or the file cannot be saved successfully, then error –25x is returned. If the save is successful, then advisory 0.1500 is shown.

No

(Trace 1|Trace 2|Trace 3|Trace 4|Trace 5|Trace 6)( with|without) headers

Trace 1

Selects the Trace 1 register as the destination for the imported data

Key Path

Mode

Save, Data (Export), Trace

VSA, LTE, LTETDD, IDEN

Trace 2

Selects the Trace 2 register as the destination for the imported data

Key Path

Save, Data (Export), Trace

414

Common Measurement Functions 2

Mode VSA, LTE, LTETDD, IDEN

Trace 3

Selects the Trace 3 register as the destination for the imported data

Key Path

Mode

Save, Data (Export), Trace

VSA, LTE, LTETDD, IDEN

Trace 4

Selects the Trace 4 register as the destination for the imported data

Key Path

Mode

Save, Data (Export), Trace

VSA, LTE, LTETDD, IDEN

Trace 5

Selects the Trace 5 register as the destination for the imported data

Key Path

Mode

Save, Data (Export), Trace

VSA, LTE, LTETDD, IDEN

Trace 6

Selects the Trace 6 register as the destination for the imported data

Key Path

Mode

Save, Data (Export), Trace

VSA, LTE, LTETDD, IDEN

Include Header

Allows you to select whether or not the saved trace data includes header information describing scaling, formatting, etc.

Key Path

Mode

State Saved

Save, Data (Export), Trace

VSA, LTE, LTETDD, IDEN

No

Save As . . .

Key Path

Mode

Save, Data (Export)

VSA, LTE, LTETDD, IDEN

415

Common Measurement Functions 2

Screen Image . . .

Key Path

Mode

Save

VSA, LTE, LTETDD, IDEN

SPAN X Scale

This menu has softkeys for selecting measurement span and also for scaling of the X axis.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Front Panel

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Span

This controls the frequency span of the measurement. This is the full span that is displayed on a spectrum display. The actual IF bandwidth that the time record sees is 1.28 times the span. See the FREQ Channel section for details on how this interacts with start, stop, and center frequencies.

Key Path

Mode

Remote Command

Example

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

SPAN/XScale

VSA, IDEN

[:SENSe]:FREQuency:SPAN <freq>

[:SENSe]:FREQuency:SPAN?

FREQ:SPAN 10 MHZ

FREQ:SPAN?

Start Freq and Stop Freq. See narrative under FREQ Channel heading for details.

depends on span option

Saved in instrument state.

2 Hz depends on span option

Prior to A.02.00

A.02.00

Full Span

This immediate action key changes the span to the maximum available. The center frequency remains

416

Common Measurement Functions 2

unchanged, regardless of whether the Frequency Annotation property is Start/Stop or Center/Span.

Key Path

Mode

Remote Command

Example

Notes

Couplings

Readback Text

Initial S/W Revision

Modified at S/W Revision

SPAN/XScale

VSA, IDEN

[:SENSe]:FREQuency:SPAN:FULL

FREQ:SPAN:FULL

The label on the softkey gives the full span available, which depends on span option.

Changes span to maximum while keeping the center frequency constant. Start and Stop frequency are affected

[25 MHz] If playing back a recording, list the recorded bandwidth here

Prior to A.02.00

A.02.00

Signal Track

From the point of view of a spectrum display, this function, when turned on, attempts to keep the largest magnitude signal in the center of the screen. It is the equivalent of manually doing a single acquisition, doing a marker to peak search on a spectrum trace, then copying the marker position to the center frequency and repeating. (It is not necessary to be viewing a spectrum display for this function to work.)

Key Path

Mode

Remote Command

Example

Couplings

Preset

State Saved

Range

Initial S/W Revision

SPAN/XScale

VSA

[:SENSe]:VECTor|ADEMod:FREQuency:CENTer:TRACk

OFF|ON|0|1

[:SENSe]:VECTor|ADEMod:FREQuency:CENTer:TRACk?

VECT:FREQ:CENT:TRAC ON

VECT:FREQ:CENT:TRAC?

Unavailable if averaging is turned on.

0

Saved in instrument state.

On | Off

Prior to A.02.00

X Axis Scaling

By default, the X axis of a trace is scaled to show all the available data in the trace. (The exception is that in spectrum displays, the edges of the spectrum that may contain aliases are not shown by default.)

However, the X axis can be manually scaled in order to zoom in on a subset of the X values, or to set the

X scaling to more convenient numbers. X scaling may be changed even when a measurement is paused

417

Common Measurement Functions 2

or completed, and the display will be updated immediately, using the existing trace data. No measurement parameters are affected and no new measurement is made. X scaling is unique to each trace.

Scaling is based on a reference position, which may be on the left of the grid, in the center, or on the right. The X reference value is assigned to this position. The X Width is the difference between the X value on the right side of the grid and the X value on the left. If the reference is in the center, the right and left are half of the X width away.

If X scaling is set such that the left or right axis boundary falls outside the X range of the available data, the trace is shown correctly on that portion of the display where it belongs.

For Vector displays (I-Q and Constellation) the X axis is actually perpendicular to the screen, and the screen's horizontal axis is used for the real part of the Y values. In this case, the X scaling can still be used to only display a portion of the data. In the case of the X reference position, left means the least positive or most negative X value, and right means the most positive or least negative value. For example, when looking at a 10 ms time record of a QPSK signal, you could set the X reference position to left, the X reference value to 4 ms and the X width to 1 ms in order to see just the portion of the signal between 4 and 5 ms. This same portion would be shown if IQ format were chosen (even though the time axis is not visible in this case).

For Symbol tables, which are not graphed but displayed in alphanumerics, X scaling can also be used to display a portion of the complete data. For example, you can set the X reference position to left, the X reference value to 20 symbols, and the X width to 10 symbols to see symbols 20 through 30. If then change the X reference position to center, you will see symbols 15 through 25, and if you change the X reference position to right you will see symbols 10 through 20.

Annotation for the X axis is just below the grid on the left and right side. It is based on whether the X

Scaling is Auto or Man. If it is Auto, then the left side is annotated with either "Center" or "Start", and the right side is annotated with either "Span" or "Stop" followed by the appropriate numbers and units.

The Center/ Span pair is only used for Spectrum or PSD traces, and only if the Freq Annotation property is Center/Span (see Freq Annotation under the FREQ key).

If X Scaling is Man, the annotation for the left side is "Left|Ctr|Right <x_reference_value> <unit>"

(depending on the X reference position), and on the right side the annotation is "Width <x_width>

<unit>". Shown below is an illustration of two of these manual X scale annotations:

418

Common Measurement Functions 2

For Vector displays, the X axis annotation is replaced by annotation for the real part of the Y value, each annotation consisting of number followed by a unit (usually volts).

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

SPAN X Scale

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Select Trace

This function is a duplicate of the same function found on the Trace/Detector menu. See the description there for details. It is placed here to allow you to conveniently choose which trace the X scaling applies.

Key Path

Mode

Initial S/W Revision

SPAN X Scale, X Axis Scaling

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

419

Common Measurement Functions 2

Modified at S/W Revision A.02.00

X Scale

If this function is set to Auto, it causes the trace to display all available trace data. (Exception: the display of the outer edges of a spectrum which may contain aliases is governed by the All Frequency Points function setting – see below.). The annotation is updated as needed, but the X Reference Value and X

Width keys are grayed out and not updated. When this function is set to Man, the X Reference Value and

X Width softkey readbacks are updated with the current values.

Key Path

Mode

Measurement

Remote Command

Example

Couplings

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

SPAN/XScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:COUPle

OFF|ON|0|1

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:COUPle?

:DISP:VECT:TRAC1:X:COUP ON

DISP:VECT:TRAC1:X:COUP?

Forced to Man if X Reference Value or X Width is set by user.

1

Saved in instrument state.

Auto | Man

Prior to A.02.00

A.02.00

X Reference Value

This function controls the X value of the selected trace at the chosen X Reference Position (see below). It has no effect on hardware input settings.

Key Path

Mode

Measurement

Remote Command

Example

SPAN/XScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:RLEVel <real>

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:RLEVel?

DISP:VECT:TRAC:X:RLEV 1e9

DISP:VECT:TRAC:X:RLEV?

420

Common Measurement Functions 2

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

If X Scale is set to Auto, the X Reference Value is determined by the trace data and this key is grayed out.

Depends on trace

Saved in instrument state.

–9.9E+37

9.9E+37

Prior to A.02.00

A.02.00

X Width

This sets the width of the X axis that is displayed for the selected trace. The X width may be set less than the Span for frequency-domain traces, allowing you to zoom in on just a portion of the measured values.

Likewise it may be less than time span covered by time-domain data This plus the X Reference Value and X Reference Position control the range of X values that may be displayed on a trace. For example, if the X Reference position is Center, the X Reference value is 1 GHz and the X Width is 20 MHz.

Key Path

Mode

Measurement

Remote Command

Example

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

SPAN/XScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:SPAN <real>

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:SPAN?

DISP:VECT:TRAC:X:SPAN 10e6

DISP:VECT:TRAC:X:SPAN?

If X Scale is set to Auto, the X Width is determined by the trace data and this key is grayed out.

Depends on trace

Saved in instrument state.

–9.9E+37

9.9E+37

Prior to A.02.00

A.02.00

X Reference Position

This determines the position from which the X scaling is calculated for the selected trace. It may be set to

421

Common Measurement Functions 2

the left side, center, or right side of the grid.

Key Path

Mode

Measurement

Remote Command

Example

Couplings

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

SPAN/XScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:RPOSition

LEFT|CENTer|RIGHt

:DISPlay:<meas>:TRACe[1]|2|3|4:X[:SCALe]:RPOSition?

DISP:VECT:TRAC1:X:RPOS LEFT

DISP:VECT:TRAC1:X:RPOS?

If X Scale is set to Auto, the X Reference Position is determined by the trace data and this key is grayed out.

CENT

Saved in instrument state.

Left|Ctr|Right

Prior to A.02.00

A.02.00

All Frequency Points

Spectrum trace data (and PSD) are based on the FFT algorithm. By default, the outer edges of the spectrum are not displayed because they may show spurious results that are aliases of real signals that are not completely filtered out by the IF filter. For example, in the case of a 1024 point FFT only 801 points are displayed. If you want to view the additional FFT points at the edges of spectral displays, turn this function on. It is global to all traces, not specific to a single trace.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Couplings

Preset

SPAN/XScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:AFPoints OFF|ON|0|1

:DISPlay:<meas>:AFPoints?

DISP:VECT:AFP ON

DISP:VECT:AFP?

ac

Only applies if trace is showing Spectrum or PSD results.

OFF

422

Common Measurement Functions 2

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Saved in instrument state.

On | Off

Prior to A.02.00

A.02.00

Freq Annotation

This controls how Spectrum and PSD traces are annotated when their X Scale is set to Auto. If Freq

Annotation is set to Center/Span, the X-axes on windows containing frequency domain traces are labeled with the center frequency on the left and the span on the right. If the Freq Annotation is set to Start/Stop, then the start and stop frequencies appear in place of center and span. If the X Scale is manual, then this annotation style does not apply.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

SPAN/XScale

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:FANNotation CSPan|SSTop

:DISPlay:<meas>:FANNotation?

DISP:VECT:FANN CSP

DISP:VECT:FANN?

CSP

Saved in instrument state.

Center/Span | Start/Stop

Prior to A.02.00

A.02.00

Copy X Scale

This front-panel only function copies the following X scaling information from the selected trace to another:

• X reference Position

• X Reference Value

423

Common Measurement Functions 2

• X Width

• X Scale (Auto/Man)

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

SPAN X Scale, X Axis Scaling

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Sweep/Control

This key displays a menu allowing you to control time-related measurement parameters, and to pause or resume the measurement.

Initial S/W Revision

Key Path

Mode

A.01060 or later

Front Panel

VSA

Main Time

This key allows you to control the length of the overall time record used in the measurement. Note that the Gate function (see below) allows you to analyze only a portion of the displayed Main Time. Time length and Res BW are related by the following equation:

Res BW = ENBW / T where ENBW is the normalized effective noise bandwidth of the Window (see the FFT Window topic under BW for more details).and T is the time record length (in seconds).

Therefore, if you change Main Time, the Resolution bandwidth must also change, and vice versa.

Time record size (in sample points) can vary between 16 points and the full FFT size used for spectrum calculations. The FFT size is indirectly chosen by setting Freq Points (see below) and is equal to (Freq

Points – 1)* 1.28.

Main Time length (in seconds) is the time record size times the sample period. The sample period for the

Main Time result is 1/(1.28*Span).

Limits:

The maximum Main Time length is:

Max FFT size / (1.28 * Span) = (409600)/Spanif Freq points state parameter is set to Auto

FFT size / (1.28 * Span) = (Freq Points – 1)/Spanif Freq points parameter is manually set

Note that the minimum Res BW is related to maximum Main Time length.

The minimum Main Time length is

16 points / (1.28 * Span) = 12.5/Span

424

Common Measurement Functions 2

See Res BW and Res BW Coupling sections for details on couplings that can change Main Time length due to Res BW changes.

Remote Command

Example

Initial S/W Revision

Key Path

Mode

Measurement

Notes

Couplings

Preset

State Saved

Min

Max

[:SENSe]:<meas>:SWEep:TIME <time>

[:SENSe]:<meas>:SWEep:TIME?

VECT:SWE:TIME 3 MS

VECT:SWE:TIME?

A.01060 or later

Sweep Control

VSA

<meas>:=VECTor|ADEMod

This key is not available in measurements other than Vector or Analog

Demod. The annotation is shown, however. In other measurements the time length is determined by number of symbols.

Affected by Res BW, Span, Freq Points, and Window. See Res BW and Res

BW coupling sections for details.

12.75e–6

Saved in instrument state.

–9.9e37

9.9e37

Pause / Resume

Pauses or resumes acquisition at the end of the current time record acquisition.

For more information see the Measurement Functions, Sweep/Control for a description of this function.

Initial S/W Revision

Key Path

Mode

A.01060 or later

Sweep Control

VSA

Gate

This key accesses a menu of time gating control functions. Time gating lets you isolate a portion of a

Main Time record to be used for downstream spectrum and statistical analysis (instead of the whole time record). The gate position may be changed during a stopped measurement and the instantaneous gate time and spectrum traces update immediately. Averages are restarted when gate properties change. The windowing function used in gated measurements is the same as non-gated measurements.

For more information see the Measurement Functions, Sweep/Control, Gate for a description of this

425

Common Measurement Functions 2

function.

Initial S/W Revision

Key Path

Mode

A.01060 or later

Sweep Control

VSA

Gate

This Boolean softkey turns time gating on or off

Remote Command

Example

Initial S/W Revision

Key Path

Mode

Measurement

Preset

State Saved

[:SENSe]:<meas>:SWEep:EGATe:STATe OFF|ON|0|1

[:SENSe]:<meas>:SWEep:EGATe:STATe?

VECT:SWE:EGAT:STAT ON

VECT:SWE:EGAT:STAT?

A.01060 or later

Sweep/Control

VSA

<meas>:=VECTor|ADEMod

0

Saved in instrument state.

Gate Length

This adjusts the time between the beginning and the end of the gate.

Remote Command

Example

Initial S/W Revision

Key Path

Mode

Measurement

Couplings

Preset

State Saved

[:SENSe]:<meas>:SWEep:EGATe[:SPAN] <time>

[:SENSe]:<meas>:SWEep:EGATe[:SPAN]?

VECT:SWE:EGAT 2 MS

VECT:SWE:EGAT?

A.01060 or later

Sweep/Control

VSA

<meas>:=VECTor|ADEMod

Gate length and delay are limited so that the gate always falls within the current time record. If the time record length decreases, the gate delay is limited first, then the gate length.

1.28125e–6

Saved in instrument state.

426

Common Measurement Functions 2

Min

Max

16 time samples

Time record length

Gate Delay

This adjusts the time between the start of the time record and the beginning of the gate .

Remote Command

Example

Initial S/W Revision

Key Path

Mode

Measurement

Couplings

Preset

State Saved

Min

Max

[:SENSe]:<meas>:SWEep:EGATe:DELay <time>

[:SENSe]:<meas>:SWEep:EGATe:DELay?

VECT:SWE:EGAT:DEL 500 US

VECT:SWE:EGAT:DEL?

A.01060 or later

Sweep/Control

VSA

<meas>:=VECTor|ADEMod

Gate length and delay are limited so that the gate always falls within the current time record. If the time record length decreases, the gate delay is limited first, then the gate length.

0

Saved in instrument state.

0

Time record length – gate length

Freq Points

By default, the analyzer chooses the number of Freq Points displayed in Spectrum or PSD displays, depending on the Res BW or Main Time length chosen. This softkey allows you to manually enter the number of displayed frequency points. Auto mode is recommended. The number of Freq Points is related to the number of FFT points used in spectrum calculations (which is always a power of 2).

Freq Points = (FFT points)/1.28 + 1

Note that if All Frequency Points is turned on for a selected trace, then all computed FFT points are shown. (See SPAN/X scale, All Frequency Points.)

Remote Command

[:SENSe]:<meas>:SWEep:POINts <integer>

[:SENSe]:<meas>:SWEep:POINts?

[:SENSe]:<meas>:SWEep:POINts:AUTO OFF|ON|0|1

[:SENSe]:<meas>:SWEep:POINts:AUTO?

427

Common Measurement Functions 2

Example

Initial S/W Revision

Key Path

Mode

Measurement

Notes

Couplings

Preset

State Saved

Range

VECT:SWE:POIN 801

VECT:SWE:POIN?

VECT:SWE:POIN:AUTO ON

VECT:SWE:POIN:AUTO?

A.01060 or later

Sweep Control

VSA

<meas>:=VECTor|ADEMod

Keyboard entry or setting this by SCPI forces state to manual. Any entry other than a valid value is rounded up to the next available value (or limited to the maximum).

This key is not shown in measurements other than Vector or Analog Demod.

See Res BW Coupling section

See Res BW Coupling section

801

1

Saved in instrument state.

51|101|201|401|801|1601|3201|6401|12801|25601|51201|102401|204801|4096

01

Trace/Detector

This key accesses a menu allowing you to select various trace parameters for all VSA based measurements.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Front Panel

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Select Trace

This softkey brings up a menu that allows you to select the trace that is to receive the action of all successive trace-specific commands like scaling, assignment of trace data, etc. The selected trace is outlined in green and is always made visible. While the Select Trace menu is showing, Each visible trace is annotated in the middle with its own trace number, as shown: below. The trace number annotations disappear when any other menu is showing.

428

Common Measurement Functions 2

Grid 2x2 layout showing trace annotations when Trace Select dialog is active

This softkey also appears in the X and Y scaling menus. There is only one selected trace at any time. If you change which trace is selected, that change is reflected in this softkey/menu wherever it appears.

Other ways to select a trace include use of the Next Window key, clicking within a trace window with a mouse cursor, and issuing a trace-specific SCPI command.

There is no SCPI command associated with this function. Instead, SCPI commands that are trace-specific have an index on the TRACe node that determines the selected trace. Using such a command has the side effect that the trace addressed by the SCPI command becomes the selected trace for any front panel interaction.

Key Path

Mode

Notes

Couplings

Range

Readback Text

Trace/Detector (also Span / X Scale or AMPTD ? Y Scale)

VSA, LTE, LTETDD, IDEN

No SCPI. Front panel only.

Affects any trace-specific commands

Trace 1|Trace 2|Trace 3|Trace 4|Trace 5|Trace 6

Trace <n>

429

Common Measurement Functions 2

Initial S/W Revision

Modified at S/W Revision

Prior to A.02.00

A.02.00

Data

This accesses a menu of Trace data choices for the selected trace. A VSA Measurement may produce many different results from a single scan; either a graph or a table. In addition, the ACP and OBW functions can be enabled on any trace showing a frequency-domain result, and produce Summary table results. Any of these results may be assigned to a trace and displayed. See the individual measurement

PDs for a list of all Trace results that are available for each measurement.

The following Trace Data types are available in all measurements:

Soft Key Name

No Data

Spectrum

Inst Spectrum

Raw Main Time

OBW Summary for Trace 1

OBW Summary for Trace 2

OBW Summary for Trace 3

OBW Summary for Trace 4

ACP Summary for Trace 1

ACP Summary for Trace 2

ACP Summary for Trace 3

ACP Summary for Trace 4

SCPI string form

"No Data"

"Spectrum1"

"Inst Spectrum1"

"Raw Main Time1"

"Obw Summary Trc1"

"Obw Summary Trc2"

"Obw Summary Trc3"

"Obw Summary Trc4"

"Acp Summary Trc1"

" Acp Summary Trc2"

" Acp Summary Trc3"

" Acp Summary Trc4"

The following Data Registers are also available for display if there are traces stored in them (see

Trace/Detector, Copy to Data Register and Recall, Data, Trace)

"D1", "D2", "D3", "D4", "D5", and "D6"

Key Path

Mode

Measurement

Remote Command

Trace/Detector, Data

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:FEED <string>

:DISPlay:<meas>:TRACe[1]|2|3|4:FEED?

430

Common Measurement Functions 2

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

DISP:VECT:TRAC1:FEED "Spectrum1"

DISP:VECT:TRAC1:FEED?

Depends on trace number and measurement

Saved in instrument state.

see table above and in individual Measurement PDs

Prior to A.02.00

A.02.00

The complete list of Trace Data names that may be assigned using the above SCPI can be obtained by using the following SCPI query:

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:DATA[1]|2|3|4:NAMes?

CALC:VECT:DATA:NAM?

Query only. Returns a comma-separated list of trace data names that may be used in DISPlay:<meas>:TRACe[1]|2|3|4:FEED "<string>". The list is the same regardless of trace index.

Prior to A.02.00

A.02.00

Spectrum

This key assigns the selected trace to display the Spectrum data result.

The Spectrum trace data displays the spectrum of the selected channel. The spectrum computation displays frequency on the x axis and amplitude on the y axis.

The following formulas show how the analyzer calculates spectrum information:

431

Common Measurement Functions 2

No Average rms Average rms Exponential AF[n]Average

Continuous Peak Hold Average

Time Average

Time Exponential Average

As shown in the previous formulas, the spectrum may be a linear spectrum or power spectrum as follows:

If the average is...

Averaging OFF rms Average

Continuous peak

then the spectrum is...

Linear

Power

Power

Linear spectra contain magnitude and phase (real and imaginary) information. Remember however, that scalar measurements (produced by the Spectrum Analyzer application) do not provide phase

432

Common Measurement Functions 2

information. Therefore, for scalar measurements, linear spectra have no phase information.

Power spectra contain only magnitude (real) information. This occurs with rms averages, for instance, because the results of the FFT are squared. Remember that the FFT yields both real and imaginary information. When the analyzer squares the results of the FFT, the imaginary part becomes zero.

See also Trace/Detector, “Data” on page 430

.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Inst Spectrum

This key assigns the selected trace to display the Inst. Spectrum data result.

Inst Spectrum trace data displays the instantaneous spectrum for the selected input channel.

Instantaneous spectrum is computed before data is averaged, which allows you see spectrum data before the data is averaged with other spectrum data.

NOTE

Inst Spectrum is not available when analog or digital demodulation is selected.

The following block diagram shows where, in the block diagram, spectrum and instantaneous spectrum are created.

Window and

FFT

Average

Spectrum

Inst. Spectrum

This measurement calculation is useful for these types of averaged measurements:

• rms

• rms exponential

• Continuous peak hold

If averaging is off, the spectrum and instantaneous spectrum display the same information.

See also Trace/Detector,

“Data” on page 430

.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

433

Common Measurement Functions 2

Main Time

This key assigns the selected trace to display the Main Time data result.

Note that Main Time is not available when analog or digital demodulation is selected.

Main Time versus Gate Time

The term is used to differentiate between the "main" time record and the "gate" time record when time gating is on.

A time record is the basic building block of the Fast Fourier Transform (FFT). The FFT takes the time-domain information in the time record and transforms it into the frequency domain.

When time gating is on, you can identify a portion of the main time-record to be used by the FFT. The term "main time-record" identifies the entire time record; the term "gate time-record" identifies the portion selected by the gate.

Selecting the Main Time trace data displays the entire time record--the main time-record. Selecting the Gate Time trace data displays that portion of the main time-record marked by the gate--the gate time-record.

The following block diagram shows the blocks that create main time and gate time.

Note that the Analog Demodulation block is available only when analog demodulation is enabled

Instantaneous

Main Time

Main Time

Gate Time

Ch1

Ch2

Time

Corrections

Time

Average

Time

Gating

Window and

FFT

Analog

Demod

There are many reasons why you may want to view the main time record. Here are just a few:

To verify that there is an input signal.

To see the characteristics of the input signal.

To help in manually setting the input range.

Time Records and Span

If you set the analyzer to full span, the time data you see is the actual input time-record. This is raw input data--the signal from which all subsequent measurements are based.

If you set the instrument to measure a specific bandwidth (something less than full span), the time data you see is the raw input data after it has been filtered (to provide alias protection) and decimated (to obtain the desired span).

Time Records and Averaging

If rms or continuous peak-hold averaging is on, the analyzer displays the most recent time record. The analyzer does not show an averaged time waveform, because all averaging is done after the time data has been transformed to the frequency domain.

If time averaging is on, the analyzer displays the averaged time-record. In other words, the time record has been averaged with previous time records.

How the Analyzer Displays the Time Record

434

Common Measurement Functions 2

It is important to remember that although the time record looks like an oscilloscope display, the analyzer is not a digital oscilloscope.

The time record represents samples of a waveform. The samples have enough information to accurately reconstruct the input signal--but the human eye may not properly perform the reconstruction. In fact, for frequencies that are higher than about ten percent of the frequency span, there will be noticeable visible distortion.

The analyzer's anti-alias filters will cause some ringing or distortion of square waves or transients when viewed in the time domain.

See also Trace/Detector, “Data” on page 430 .

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data

VSA

Prior to A.02.00

A.02.00

Inst Main Time

This key assigns the selected trace to display the Inst Main Time data result.

Inst Main Time trace data displays the instantaneous time-domain data for the selected input channel.

Note that Inst Main Time is not available when analog or digital demodulation is selected.

The following block diagram shows how Instantaneous Main Time is derived.

Instantaneous

Main Time

Main Time

Gate Time

Ch1

Ch2

Time

Corrections

Time

Average

Time

Gating

Window and

FFT

Analog

Demod

Notice that Instantaneous Main Time shows you time data before time averaging. If time averaging is off, Instantaneous Main Time is identical to Main Time.

See also Trace/Detector,

“Data” on page 430

.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

435

Common Measurement Functions 2

Gate Time

This key assigns the selected trace to display the Gate Time data result.

Gate Time trace data displays the selected channel's gate time-record.

Note that Gate Time is not available when analog or digital demodulation is selected.

If time gating is on, Gate Time displays the portion of the main time-record marked by the gate-- this portion is called the gate record (if time gating is off, Gate Time displays nothing).

As a reminder, if time gating is on, the Fast Fourier Transform (FFT) uses the gate time-record, which can be all or a portion of the main time-record, to compute frequency information such as spectrum, frequency response, coherence, and correlation.

See also Trace/Detector, “Data” on page 430 .

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Raw Main Time

This key assigns the selected trace to display the Raw Main Time data result.

Raw Main Time is the raw data read from the input hardware or playback file. It is similar to

“Raw Main

Time” on page 436

with the following exceptions:

• This data has not had time corrections applied, so it displays a "CAL?" trace indicator.

• The data has not gone through the analyzer's software resampling filters, so is generally not sampled at the specified sample rate.

• The data has a wider bandwidth than the measurement span would indicate.

Raw Main Time data is useful in the following situations:

• When you use Channel, IF Magnitude, or Magnitude trigger types, the input hardware detects the trigger, so Raw Main Time sometimes gives a better indication of what caused the trigger.

• When you play back a recording, the Raw Main Time measurement data allows you to see exactly the samples that are saved in the recording, with no filtering applied or settling removed.

See also Trace/Detector, “Data” on page 430 .

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

436

Common Measurement Functions 2

PSD (Power Spectral Density)

This key assigns the selected trace to display the Power Spectral Density (PSD) data result.

PSD trace data displays the power spectral density (PSD) of the selected channel.

The definition of PSD yields y-axis units of Vpk2/Hz and x-axis units of frequency:

PSD is used for noise measurements. It shows the power density of a signal as a function of frequency. In general, noise may have any arbitrary frequency content, resulting in a variety of possible PSD shapes.

Noise that has equal power density at all frequencies is called white noise:

The definition of PSD is power per Hertz. In other words, power is divided by the measurement bandwidth, which in this analyzer is the resolution bandwidth (ResBW), as follows:

Units of Vpk2/Hz assumes the signal is referenced to 1 ohm. That is, because no resistance is specified, the signal is interpreted as a voltage across a one ohm resistor, with the power in the resistor equal to

Vpk2.

You can select units of dBm/Hz to take into account the analyzer's input impedance. PSD defaults to these units. The analyzer calculates dBm/Hz as follows:

437

Common Measurement Functions 2

See also Trace/Detector, “Data” on page 430 .

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Auto Correlation

This key assigns the selected trace to display the Power Spectral Density (PSD) data result.

Auto Correlation trace data displays the autocorrelation for the selected input channel. Autocorrelation is a form of correlation, a measure of the similarity between two signals.

Note that Auto Correlation is not available when digital demodulation is selected.

Tips

Use ac coupling only. Correlation measurements are disturbed by dc offsets in the signal.

Some types of averaging may be useful -- rms averaging does not affect correlation measurements, but you can use time averaging to reduce noise, if you can provide a consistent trigger. However, averaging is usually unnecessary to make good correlation measurements.

Use appropriate triggering and trigger delays. This is especially true for time averaging.

Use a random noise source for delay measurements. Correlation measurements provide the ability to resolve time differences between waveforms that appear to be random.

Waveforms on the correlation trace may not appear as they do in the time trace. This is particularly noticeable when you are using correlation to extract synchronous signals from noise. The different shape of some waveforms is a direct result of the mathematical definition of correlation. For example, a correlated square wave appears as a triangle wave. It's important to remember that the period of the waveform is preserved even if the correlation waveform looks different.

To avoid wrap-around effects, correlation produces a time record one-half the length of the measurement time-record.

Theory of Operation

Autocorrelation is a form of correlation, a measure of the similarity between two signals. Correlation is performed by multiplying two signals together at each instant in time and summing all the products. If the signals are identical, every product is positive and the resulting sum is large.

438

Common Measurement Functions 2

If, however, the two signals are dissimilar, then some of the products are positive and some are negative.

In this case, the final sum is smaller because the products tend to cancel.

Autocorrelation performs a time-shifted, "averaged" correlation on a single signal. The signal is correlated with time-shifted versions of itself. Furthermore, the products from each time-shift are averaged by dividing each final sum by the number of products contributing to it.

T

→∞

× τ where: Rxx = autocorrelation function

τ = amount of time shift

= infinity x(

τ) = signal to be correlated intgrl = integration conj = conjugation

T = time

× = multiplication

That is, the autocorrelation function is found by taking a signal, multiplying it by the same signal displaced (tau) units in time, and averaging the product over all time.

Duality With the Power Spectrum

For simplicity and speed, this analyzer performs the autocorrelation operation by taking advantage of its duality with the power spectrum:

Thus,

When to use Auto Correlation

Auto correlation is useful for detecting echoes in a signal. For random noise, an echo appears as an impulse -- if there is more than one echo, you will see multiple peaks on the auto correlation trace. Keep in mind that an echo appears as an impulse only if the delayed signal has not been filtered. The impulse broadens as the original random noise signal is filtered -- in fact, the width of each peak is inversely proportional to the bandwidth of the signal.

To determine the time delay (in seconds) of an echo, you can move the marker to the peak of the echo. Note that there is always a correlated peak at zero lag -- this peak marks the original excitation signal. Any other peaks let you know that the excitation signal also appeared at another time relative to the original signal. The amplitude value at the zero lag point is the total power in the time record.

This function is also useful for isolating low-level periodic signals from noise. A sine wave signal shows up as a

439

Common Measurement Functions 2

sine wave in auto correlation. A square wave signal shows up as a triangular wave of the same frequency.

Auto correlation is a single-channel measurement. If you have the original signal on one channel and the delayed version on another, use cross correlation.

Auto Correlation and Averaging

The following formulas show how the analyzer calculates auto correlation for different averaging functions:

Key: F = Fast Fourier Transform (FFT)

AC = Averaged correlation

AT = Averaged time t = Instantaneous time c = Instantaneous correlation r = 1/2 width rectangular window

× = multiplication

n = Average number

No Average c = I (conj ( F(r

× t) ) × F [t] ) rms Average c = I (conj ( F(r

× t) ) × F [t] ) rms Expon. c = I (conj ( F(r

× t) ) × F [t] )

Average

Continuous

Peak Hold c = I (conj ( F(r

× t) ) × F [t] )

Average

Time AC [n] = I (conj ( F(r

× AT [n]) ) × F (AT [n] ) )

Average where: AT [n] = sum (t [n]) n

Time

Expon. AC[n] = I (conj ( F(r

× AT [n]) ) × F (AT [n] ) )

Average where: AT [n] = t [n] + AT [n 1] n n

− and: 1 < n < number of averages

See also Trace/Detector,

“Data” on page 430

.

Key Path

Mode

Initial S/W Revision

Trace/Detector, Data

VSA

Prior to A.02.00

440

Common Measurement Functions 2

Statistical

This key accesses the Trace Data choices which show statistical results.

Key Path

Mode

Trace/Detector, Data

VSA

CCDF (Complementary, Cumulative Density Function)

This key assigns the selected trace to display the CCDF data result.

CCDF trace data displays the complementary, cumulative density function (CCDF) for the selected input channel.

The complementary, cumulative density function (CCDF) is a statistical-power calculation and can be performed only on time-domain data. As its name suggests, CCDF is the complement of CDF, and is defined as follows:

CDF(K) = Probability(x

≤ K)

CCDF(K) = Probability(x

≥ K)

CCDF provides better resolution than CDF for low probability signals, especially when log format is used for the y-axis.

The analyzer plots CCDF using units of percent (%) for the y-axis and power (dB) for the x-axis. Power on the x-axis is relative to the signal average power, so 0 dB is the average power of the signal. Therefore, a marker readout of

Trace A Marker 2 dB 12 % means there is a 12% probability that the signal power will be 2 dB or more above the average power.

CCDF Calculation:

Calculate the RMS value for all measured samples; this becomes the 0 dB point at the left end of the x-axis.

Normalize all samples to the RMS value in units of dB.

Determine which x-axis bin each sample belongs in between 0 and 20 dB.

Calculate the total number of samples that are greater than or equal to each x-axis bin and plot as a percent of the number of samples measured.

Samples Used in the Power Measurement

For the Demod Off and Analog demod modes, the analyzer computes CCDF using all samples in the current time record (all points in the active trace). Each successive time record adds additional samples to the CCDF measurement.

For WLAN - OFDM and -DSSS demod modes, the analyzer computes CCDF using all samples specified within the measurement interval.

Restarting the Power Measurement

Selecting CCDF, restarting the measurement, or changing most measurement parameters restarts the CCDF measurement. For example, changing the range or center frequency resets the number of samples used in the CCDF measurement to zero and restarts the CCDF measurement.

Tips

441

Common Measurement Functions 2

Note the following when making CCDF measurements:

For best results, set the analyzer's displayed frequency span to include all the energy of your signal. In other words, make sure the displayed frequency span includes the entire bandwidth of the measured signal.

The CCDF measurement does not restart:

After a calibration

After you continue a paused measurement

Many channel specific changes restart the CCDF measurement on both channels, such as changing the gate delay, or input coupling.

The analyzer displays DATA? if the average power drifts 8 to 10 dB from the average power measured in the first time record. For example, the analyzer would display DATA? if you measured a transmitter signal that was off when the CCDF measurement started but then turned on later in the measurement.

CCDF measurements are disabled during time averaging.

See also Trace/Detector,

“Data” on page 430

.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

CDF (Cumulative Density Function)

This key assigns the selected trace to display the CDF data result.

CDF trace data displays the Cumulative Density Function (CDF) for the selected input channel. CDF is computed by integrating the PDF (Probability Density Function). CDF is explained in online help for the CCDF trace data for further details, see that topic

See also Trace/Detector,

“Data” on page 430

.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

PDF (Probability Density Function)

This key assigns the selected trace to display the PDF data result.

PDF trace data displays the Probability Density Function of the selected channel. PDF indicates the probability that a given level has occurred.

PDF is equivalent to a normalized histogram. A histogram shows how the amplitude of a signal is distributed between its maximum and minimum values. Amplitude is displayed on the X-axis, and number of counts on the

Y-axis.

442

Common Measurement Functions 2

The number of averages for a histogram determines the number of counts in the histogram; in other words, how many records are measured¾the records are not "averaged". If averaging is off or if exponential averaging is selected, the measurement continues indefinitely. Keep in mind that the accuracy of the histogram is dependent on the frequency span, time-record length, and number of averages (if averaging is on).

Histograms are used for such things as determining the statistical properties of noise and monitoring the performance of electromechanical positioning systems.

PDF trace data is normalized by multiplying the number of averages by the number of points in the time record, then dividing this value by the DV spacing on the X-axis. The probability of a signal falling between two points is equal to the integral of the curve between those points.

PDF trace data displays the number of points used in its computation above the trace (Pts:). It also displays the average level (Avg:) above the trace.

See also Trace/Detector, “Data” on page 430 .

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Format

This accesses a menu that allows you to choose the format of the selected trace. Any format may be assigned to any trace. For symbol tables and tabular data the format choice is ignored. If the data doesn't have defined symbol times, Constellation format is the same as I-Q, Eye formats are the same as Real or

Imaginary, and Trellis format is the same as Unwrapped Phase.

The formats are:

Format name

Log Mag (dB)

Linear Mag (Abs Value)

Real (I)

Imaginary (Q)

I-Q

Constellation

Wrap Phase

Unwrap Phase

Description

Data is converted to decibel units and shown on a linear Y axis

Magnitude of the data is shown on a linear Y axis

Real part of data is shown on a linear Y axis

Imaginary part of data is shown on linear Y axis

Real part of data is shown on horizontal axis, imaginary part is shown on vertical axis, Independent variable (X axis) is normal to display

Same as I-Q, but for data with symbols defined, only the symbol points are shown as dots with no connecting lines.

Phase of complex data, limited to ± 180 deg, is shown on Y axis

Phase of complex data is shown "unwrapped", that is, without discontinuities. Not limited to ± 180 degrees.

443

Common Measurement Functions 2

Format name

I-Eye

Q-Eye

Trellis

Group Delay

Log Mag (Linear Unit)

Description

Real part of data is shown with X axis segmented (generally into 2 symbol segments) and each segment is overlaid to show signal crossings at symbol boundaries

Same as I-eye but imaginary part of data is shown

Same as I-eye but uses unwrapped phase of data

Useful for frequency response displays. Shows the derivative of phase response with respect to frequency.

Displays data with a logarithmic Y axis, but marker read outs are in linear magnitude units.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Range

Readback Text

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Format

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat

MLOG|MLINear|REAL|IMAGinary|VECTor|CONS|PHASe|UPHase|IE

YE|QEYE|TRELlis|GDELay|MLGLinear

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat?

DISP:DDEM:TRAC2:FORM MLIN

DISP:DDEM:TRAC2:FORM?

Depends on trace and measurement

Saved in instrument state.

Log Mag (dB)|Linear Mag (Abs Value)|Real (I) (Lin)|Imaginary (Q)

(Lin)|I-Q|Constellation|Wrap Phase|Unwrap

Phase|I-Eye|Q-Eye|Trellis-Eye|Group Delay|Log Mag (Linear Unit)

Log Mag (dB)|Linear Mag|Real (I)|Imaginary (Q)|I-Q|Constellation|Wrap

Phase|Unwrap Phase|I-Eye|Q-Eye|Trellis-Eye|Group Delay|Log Mag

Prior to A.02.00

A.02.00

Digital Demod Trace Setup

This key accesses a menu of settings that control certain elements of displays of digitally demodulated trace data.

Key Path

Mode

Initial S/W Revision

Trace/Detector

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

444

Common Measurement Functions 2

Modified at S/W Revision A.02.00

Symbol Shape

For all time-domain displays except IQ diagrams, Symbol Shape lets you display dots, bars, or nothing

(none) at symbol locations (if the trace contains demodulated time-domain data). This key allows you to select the symbol shape for the selected trace.

If you select bars, vertical lines (bars) are drawn from the baseline to the symbol location on the trace.

The baseline is 0 for all traces that have coordinates other than log (dB). The baseline is the bottom of the trace box for traces that have log (dB) coordinates.

With IQ diagrams, displaying vertical bars is meaningless. Therefore, selecting bars displays dots in IQ diagrams.

With constellation diagrams, selecting none is the same as selecting bars--you cannot turn off the dots in a constellation diagram.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Digital Demod Trace Setup

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol

BARS|DOTS|OFF

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol?

DISP:DDEM:TRAC2:DDEM:SYMB DOTS

DISP:DDEM:TRAC2:DDEM:SYMB?

BARS

Saved in instrument state.

Bars|Dots|None

Prior to A.02.00

A.02.00

Ideal State Shape

Digital Demodulation shows you the location of all ideal symbol states in an I-Q or constellation diagram. This key lets you choose between a cross, circle, or none to represent the ideal state on the selected trace.

Key Path

Mode

Measurement

Trace/Detector, Digital Demod Trace Setup

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

445

Common Measurement Functions 2

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol:SHAPe

CIRCle|CROSs|OFF

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol:SHAPe?

DISP:DDEM:TRAC2:DDEM:SYMB:SHAP CIRC

DISP:DDEM:TRAC2:DDEM:SYMB:SHAP?

CIRC

Saved in instrument state.

Circle|Cross|None

Prior to A.02.00

A.02.00

Ideal State Size

Determines the ideal state size, as a percentage of the maximum ideal state distance from the origin (the same way Error Vector Magnitude is defined). Ideal states are shown as circles or crosses in Vector and constellation diagrams, as determined by the Ideal State Shape setting.

The ideal state is where symbols occur if your signal is without error. Showing the ideal states gives a visual indication of the quality of your signal.

You can use this feature to determine if symbols have an EVM above a specified Value. For example, to see if any symbols have an EVM greater than 10%, set the state size to 10% and select Circle as the shape. Any symbols that fall outside of the circle (other than SYNC or PILOT symbols) have an EVM greater than 10%.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Preset

State Saved

Min

Max

Trace/Detector, Digital Demod Trace Setup

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol:SIZE

<real>

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol:SIZE?

DISP:DDEM:TRAC2:DDEM:SYMB:SIZE 10

DISP:DDEM:TRAC2:DDEM:SYMB:SIZE?

Parameter is interpreted as a percent, e.g., if you want the ideal size to be 10% send 10, not 0.1

5

Saved in instrument state.

0.1

50

446

Common Measurement Functions 2

Initial S/W Revision

Modified at S/W Revision

Prior to A.02.00

A.02.00

Symbol Table Format

This allows you to choose the format in which symbol table data is displayed, when the modulation format encodes 4 or more bits per symbol. You may choose binary or hexadecimal. Binary symbol data is padded with leading zeros to make a multiple of 4 bits before conversion to hexadecimal. For example, for 16 QAM format, each 4-bit symbol will be displayed as 2 hex digits.

Binary Format: The symbol data bit format is binary and each character represents a binary digit. The number to the left of each row indicates the bit offset of the first bit in the row.

Hexadecimal Format: The symbol data bit format is hexadecimal and each character represents a hexadecimal digit. The number to the left of each row indicate the symbol offset of the first symbol in the row.

NOTE

There must be at least 4 bits/symbol to use the hexadecimal format, that is, symbols that have less than 4 bits/symbol will only be displayed in binary format regardless of the Symbol Table Format setting.

This parameter is valid only when:

• The active trace is a symbol table, and

• The current demodulation format supports hexadecimal, the demodulation format's bits/symbol is equal to or greater than four.

Key Path

Mode

Measurement

Remote Command

Example

Preset

Range

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Digital Demod Trace Setup

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol:FORMat

HEXadecimal|BINary

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:SYMBol:FORMat?

DISP:DDEM:TRAC2:DDEM:SYMB:FORM BIN

DISP:DDEM:TRAC2:DDEM:SYMB:FORM?

HEX

Hex | Binary

Prior to A.02.00

A.02.00

Eye Length

This property controls how wide (in symbol periods) the eye and trellis diagrams are, for the selected

447

Common Measurement Functions 2

trace.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Digital Demod Trace Setup

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:EYE:COUNt <real>

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:EYE:COUNt?

DISP:DDEM:TRAC2:DDEM:EYE:COUN 3

DISP:DDEM:TRAC2:DDEM:EYE:COUN?

2

Saved in instrument state.

0.1

40

Prior to A.02.00

A.02.00

Time Unit

This property lets you select the time units that are applied to x-axis annotations and marker readouts for the selected trace, whenever it is assigned data with (demodulation) symbol information. The available measurement units are sym (symbols) or sec (seconds).

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Digital Demod Trace Setup

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:UNIT:TIME

SEC|SYMBol

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:UNIT:TIME?

DISP:VECT:TRAC2:DDEM:UNIT:TIME SYMB

DISP:VECT:TRAC2:DDEM:UNIT:TIME?

SYMB

Saved in instrument state.

sym|sec

Prior to A.02.00

A.02.00

448

Common Measurement Functions 2

Freq Unit

This property lets you select the frequency units that are applied to x-axis annotations and marker readouts for the selected trace, whenever it is assigned data with (demodulation) carrier information. The available measurement units are carrier or Hz.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Digital Demod Trace Setup

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:UNIT:FREQuency

CARRier|HZ

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:UNIT:FREQuency?

DISP:VECT:TRAC2:DDEM:UNIT:FREQ CARR

DISP:VECT:TRAC2:DDEM:UNIT:FREQ?

CARR

Saved in instrument state.

carrier|Hz

Prior to A.02.00

A.02.00

Avg Line

This controls whether or not the average line is visible on certain demodulation analysis traces such as

Error Vector Time and Error Vector Spectrum in OFDM related modulation analysis measurements.

These traces have 2-dimensional domains; typically subcarriers (frequency) and symbol times. Since the result can only be shown with one of these dimensions on the x-axis, the other dimension is placed on the z-axis. Since all the z-axis values are overlapped, an average is calculated for all z values at each x value and the average is normally displayed as a line in front of trace. The average line display can be turned on or off using this control.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Trace/Detector, Digital Demod Trace Setup

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:ALINe OFF|ON|0|1

:DISPlay:<meas>:TRACe[1]|2|3|4:DDEMod:ALIN?

DISP:W11A:TRAC:DDEM:ALIN OFF

1

Saved in instrument state.

449

Common Measurement Functions 2

Initial S/W Revision A.03.00 or later

Copy to Data Register

This key accesses a menu of immediate execute keys, each of which copies the selected trace to a particular data register. Data registers can be displayed in any trace. They are measurement global, so you can copy data to a register while in the Digital Demod measurement and view it later while in the

Vector measurement. Data registers are cleared when the VSA Application is exited and reentered, but not when you change Modes and return.

Key Path

Mode

Measurement

Remote Command

Example

Readback Text

Initial S/W Revision

Modified at S/W Revision

Trace/Detector

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:COPY D1|D2|D3|D4|D5|D6

DISP:VECT:TRAC:COPY D1

Last: <date_time>|Empty

Prior to A.02.00

A.02.00

The following SCPI provides means to determine if a Data Register is empty, and to erase the data from any or all Data Registers.

Key Path

Mode

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

SCPI only

VSA, LTE, LTETDD, IDEN

:CALCulate:DATA:REGister[1]|2|3|4|5|6:EMPTy?

:CALC:DATA:REG2:EMPT?

Query only: returns 1 if a Data Register has no trace data assigned to it.

Prior to A.02.00

A.02.00

Key Path

Mode

Remote Command

Example

Notes

Couplings

SCPI only

VSA, LTE, LTETDD, IDEN

:CALCulate:DATA:REGister[1]|2|3|4|5|6:REMove

:CALC:DATA:REG2:REM

Removes trace data assigned to specified Data Register.

If Data Register is assigned to a trace, the trace data is changed to No Data

450

Common Measurement Functions 2

Initial S/W Revision

Modified at S/W Revision

Prior to A.02.00

A.02.00

Key Path

Mode

Remote Command

Example

Notes

Couplings

Initial S/W Revision

Modified at S/W Revision

SCPI only

VSA, LTE, LTETDD, IDEN

:CALCulate:DATA:REGister:ALL:REMove

:CALC:DATA:REG:ALL:REM

Removes trace data assigned to all Data Registers.

If Data Register is assigned to a trace, the trace data is changed to No Data

Prior to A.02.00

A.02.00

Phase/Delay Properties

This key accesses a menu of properties that affect the selected trace when displayed using phase or delay formats

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Phase/Trellis Offset

This is only used if the trace Format is Wrap Phase, Unwrap Phase or Trellis. For Unwrap Phase or

Trellis displays, the phase offset value is added to the existing phase at each point. For example, If you are viewing an Unwrapped Phase trace, setting the Phase/Trellis Offset to 5 degrees moves the entire trace up 5 degrees (and changes the value displayed by a marker by the same amount). For Wrap Phase displays the phase offset only affects the phase wrap point, not the underlying data. The point at which the phase wraps is 180 degrees plus the phase offset. For example, suppose you have a marker on a Wrap

Phase display whose phase offset is 0 and the marker is showing –3 degrees. The trace data will all be confined within (–180, 180] degrees. If you then change the phase offset to 180 degrees, then the Wrap

Phase display will show values within the interval (0, 360] degrees and the marker value will be displayed as 357 degrees, which is the wrapped equivalent of –3 degrees.

Key Path

Mode

Measurement

Trace/Detector, Phase Delay Properties

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

451

Common Measurement Functions 2

Remote Command

Example

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat:PHASe:OFFSet

<real>

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat:PHASe:OFFSet?

DISP:DDEM:TRAC3:FORM:PHAS:OFFS 31

DISP:DDEM:TRAC3:FORM:PHAS:OFFS?

0

Saved in instrument state.

–1E+8

1E+8

Prior to A.02.00

A.02.00

Unwrap Phase Ref

Unwrapped phase lets you designate the point (x-axis) value about which phase values are to be unwrapped. That is, the phase at the designated reference will be within –180 to 180 degrees, and phase will vary smoothly without jumps around that point.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Phase Delay Properties

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat:PHASe:UNWRap:REFe rence <real>

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat:PHASe:UNWRap:REFe rence?

DISP:DDEM:TRAC3:FORM:PHAS:UNWR:REF 24.5E6

DISP:DDEM:TRAC3:FORM:PHAS:UNWR:REF?

0

Saved in instrument state.

–9.9e37

9.9e37

Prior to A.02.00

A.02.00

Group Delay Aperture

The value of Delay Aperture is used when the trace format is Group Delay. The aperture is specified as a percentage of the current frequency span for frequency-domain data. It is specified as a percentage of the

452

Common Measurement Functions 2

time-record length for time-domain data.

When group delay is calculated for a given point (which can be a time- or frequency-domain point), the aperture is centered at that point. Larger apertures decrease resolution, but they increase the smoothing of the group-delay trace.

The point plotted for group delay is located between the data points used to calculate it. For example, in the frequency domain, the group delay for 100 Hz may be calculated by measuring the change in phase between 90 and 110 Hz. If you had specified a start frequency of 90 Hz, 100 Hz would be the first point with group delay data. This results in a trace that does not extend to the edges of the screen (more noticeable as the delay aperture increases).

Note that the smallest aperture that you can select depends on the number of frequency points. If you select an invalid aperture, the analyzer automatically selects the smallest valid aperture.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Phase Delay Properties

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat:DELay:APERture

<real>

:DISPlay:<meas>:TRACe[1]|2|3|4:FORMat:DELay:APERture?

DISP:DDEM:TRAC3:FORM:DEL:APER 1

DISP:DDEM:TRAC3:FORM:DEL:APER?

Parameter is interpreted as a percent, e.g., if you want the group delay aperture to be 1% send 1, not 0.01

0.5

Saved in instrument state.

0.00390625

16

Prior to A.02.00

A.02.00

ACP Setup

The adjacent channel power (ACP) function calculates the power in a reference band of frequencies as well as bands of frequencies offset from the reference, and calculates the ratio of each offset band to the reference band power.

The ACP key accesses a menu of functions that allow you to define and turn on the ACP function on the selected trace. One reference channel and up to 5 offset frequencies may be defined, and ACP will be calculated for bands both above and below the reference frequency for each offset.

An ACP measurement may be defined for each trace, although it will only be active on frequency-domain trace data. The reference and offset frequency bands defined by the ACP

453

Common Measurement Functions 2

measurement are shown as gold bars overlaying the trace display. To see tabular data showing power and power ratio results, you may assign the ACP Summary (Trace n) to a different trace. For example, you can assign Spectrum data to trace 1, turn on and define an ACP measurement on trace 1, assign the ACP

Summary (Trace 1) to trace 2, and use a 2x2 display to view both at the same time, as shown below

The summary data may be retrieved programmatically using FETCh? or the

CALCulate:<meas>:DATA:TABLe commands. See the Data Queries section under Common Functions for more details.

Key Path

Mode

Readback Text

Initial S/W Revision

Modified at S/W Revision

Trace/Detector

VSA, LTE, LTETDD, IDEN

[On|Off,]

Prior to A.02.00

A.02.00

ACP Summary for Trace 1

Select trace for assignment of ACP Data.

454

Common Measurement Functions 2

See also “ACP Setup” on page 453

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data, ACP, Channel n

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

ACP Summary for Trace 2

Select trace for assignment of ACP Data.

See also “ACP Setup” on page 453

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data, ACP, Channel n

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

ACP Summary for Trace 3

Select trace for assignment of ACP Data.

See also “ACP Setup” on page 453

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data, ACP, Channel n

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

ACP Summary for Trace 4

Select trace for assignment of ACP Data.

See also “ACP Setup” on page 453

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data, ACP, Channel n

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

ACP Summary for Trace 5

Select trace for assignment of ACP Data

455

Common Measurement Functions 2

See also

“ACP Setup” on page 453 .

Key Path

Mode

Trace/Detector, Data, ACP, Channel n

VSA

ACP Summary for Trace 6

Select trace for assignment of ACP Data.

See also

“ACP Setup” on page 453

Key Path

Mode

Trace/Detector, Data, ACP, Channel n

VSA

ACP On/Off

This softkey turns the ACP function on or off for the selected trace

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, ACP

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:STATe

OFF|ON|0|1

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:STATe?

CALC:VECT:TRAC1:ACP:STATE ON

CALC:VECT:TRAC1:ACP:STATE?

0

Saved in instrument state.

Prior to A.02.00

A.02.00

Carrier Freq

This key allows you to enter the carrier frequency of the reference channel for the ACP measurement.

The carrier frequency is relative to the center frequency of the measurement. There is only one available reference carrier.

Key Path

Mode

Measurement

Trace/Detector, ACP

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

456

Common Measurement Functions 2

Remote Command

Example

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:FREQue ncy <freq>

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:FREQue ncy?

CALC:VECT:TRAC1:ACP:CARR:FREQ 100 KHZ

CALC:VECT:TRAC1:ACP:CARR:FREQ?

0

Saved in instrument state.

–9.9e37

9.9e37

Prior to A.02.00

A.02.00

Carrier Meas Noise BW

This key allows you to define the measurement noise bandwidth of the reference channel.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, ACP

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:BANDwi dth|BWIDth:INTegration <bandwidth>

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:BANDwi dth|BWIDth:INTegration?

CALC:VECT:TRAC1:ACP:CARR:BAND:INT 1 MHZ

CALC:VECT:TRAC1:ACP:CARR:BAND:INT?

1000000

Saved in instrument state.

–9.9e37

9.9e37

Prior to A.02.00

A.02.00

457

Common Measurement Functions 2

Carrier RRC Weighting

This key turns on or off RRC weighting for the reference (carrier) power measurement.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, ACP

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:FILTer

:RRC:STATe OFF|ON|0|1

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:FILTer

:RRC:STATe?

CALC:VECT:TRAC1:ACP:CARR:FILT:RRC:STAT ON

CALC:VECT:TRAC1:ACP:CARR:FILT:RRC:STAT?

0

Saved in instrument state.

Prior to A.02.00

A.02.00

Carrier Filter Alpha

This key allows you to adjust the alpha of the RRC filter for the reference (carrier) power measurement.

Key Path

Mode

Measurement

Remote Command

Example

Preset

State Saved

Min

Max

Initial S/W Revision

Trace/Detector, ACP

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:FILTer

:RRC:ALPHa <real>

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:CARRier:FILTer

:RRC:ALPHa?

CALC:VECT:TRAC1:ACP:CARR:FILT:RRC:ALPH 0.22

CALC:VECT:TRAC1:ACP:CARR:FILT:RRC:ALPH?

0

1

0.35

Saved in instrument state.

Prior to A.02.00

458

Common Measurement Functions 2

Modified at S/W Revision A.02.00

Offsets

The ACP measurement compares power in frequency bands offset from the carrier to power in the reference channel (centered on the carrier). Up to 5 offsets may be defined. The offsets are designated by letters A through E. Each offset is defined by an offset frequency, bandwidth, and optional RRC weighting. An offset actually defines two bands, one above the reference frequency and one below. Each band is used individually in the ACP calculation. RRC weighting may only be turned on or off for all offsets, but each offset may have its own RRC filter alpha. A filter alpha of 0 is the same as no RRC weighting.

The Offsets key accesses a menu that has a key for each offset, and also an Offset RRC weighting on/off key. Each offset key shows a summary of its current parameters. Pressing one of the Offset

A|B|C|D|E keys accesses a menu for adjusting its parameters

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector,ACP,Offsets

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Offset Freq

This key turns ACP analysis on or off for a selected offset, and sets the offset frequency (which is relative to the carrier frequency).

Key Path

Mode

Measurement

Remote Command

Example

Trace/Detector,ACP,Offsets,Offset A|B|C|D|E

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:FR

EQuency <freq>,…

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:FR

EQuency?

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:ST

ATe OFF|ON|0|1,…

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:ST

ATe?

CALC:VECT:TRAC1:ACP:OFFS:LIST:FREQ 1 MHZ,1 MHz,500 KHZ,500

KHz,1 MHZ

CALC:VECT:TRAC1:ACP:OFFS:LIST:FREQ?

:CALC:VECT:TRAC1:ACP:OFFS:LIST:STAT ON,OFF,OFF,ON,OFF

459

Common Measurement Functions 2

Notes

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

If you send fewer than 5 frequencies in the parameter list, then the remaining offsets frequencies are set to 0.

You may send a single on/off parameter, or a comma-separated list of up to 5 parameters. These enable/disable each of the Offsets in sequence. Any remaining Offsets are disabled

3000000,0,0,0,0

1,0,0,0,0

Saved in instrument state.

–9.9E+37

9.9E+37

Prior to A.02.00

A.02.00

Offset Meas Noise BW

This key allows you to set the measurement noise bandwidth for the power measurement of a selected offset band.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Trace/Detector,ACP,Offsets,Offset A|B|C|D|E

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:BA

NDwidth|BWIDth:INTegration <bandwidth>,…

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:BA

NDwidth|BWIDth:INTegration?

CALC:VECT:TRAC1:ACP:OFFS:LIST:BAND:INT 1 MHZ,2 MHZ,3

MHZ,4 MHZ,5 MHZ

CALC:VECT:TRAC1:ACP:OFFS:LIST:BAND:INT?

If you send fewer than 5 bandwidth parameters in the list, then Measurement

Noise Bandwidths for the remaining Offsets are set to 0.

1000000,0,0,0,0

Saved in instrument state.

–9.9e37

9.9e37

Prior to A.02.00

A.02.00

460

Common Measurement Functions 2

Offset Filter Alpha

This key allows you to adjust the alpha of the RRC filter for the power measurement of the selected offset band.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Trace/Detector,ACP,Offsets,Offset A|B|C|D|E

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:FI

LTer:RRC:ALPHa <real>,…

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:FI

LTer:RRC:ALPHa?

CALC:VECT:TRAC1:ACP:OFFS:LIST:FILT:RRC:ALPH

0.22,0.22,0.22,0.22,0.22

CALC:VECT:TRAC1:ACP:OFFS:LIST:FILT:RRC:ALPH?

You may send a single Filter Alpha for Offset A, or a comma-separated list of up to 5 Filter Alpha parameters. These are assigned in sequence to the Offsets.

Alpha for any remaining Offsets will be set to 0.

0.35,0.35,0.35,0.35,0.35

Saved in instrument state.

0

1.0

Prior to A.02.00

A.02.00

Offset Relative Limit

This key enables you to turn on/off a relative limit test and set the limit for the selected offset. The test shows a failure if the power in either the upper or lower band at the selected offset exceeds the reference power plus the relative test limit. For example, if the test limit is –60, the reference power is –4.5 dBm, a test failure would be shown if the power in the lower or upper band exceeds –64.5 dBm.

Key Path

Mode

Measurement

Trace/Detector,ACP,Offsets

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

461

Common Measurement Functions 2

Remote Command

Example

Notes

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:RC

ARrier <reall>,…

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:RC

ARrier?

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:RC

ARrier:TEST OFF|ON|0|1,…

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:LIST:RC

ARrier:TEST?

CALC:VECT:TRAC1:ACP:OFFS:LIST:RCAR –50, –55, –60, –65, –80

CALC:VECT:TRAC1:ACP:OFFS:LIST:RCAR?

CALC:VECT:TRAC1:ACP:OFFS:LIST:RCAR:TEST 1, 1, 1, 1, 1

CALC:VECT:TRAC1:ACP:OFFS:LIST:RCAR:TEST?

You may send a single Limit for Offset A, or a comma-separated list of up to 5 limit parameters. These are assigned in sequence to the Offset frequencies, with the remaining limits being set to 0.

You may send a single on/off parameter, or a comma-separated list of up to 5 parameters. These turn the Limit Test on or off for each of the Offsets in sequence. For any remaining Offsets the Limit test will be turned off.

–120,–120,–120,–120,–120

0,0,0,0,0

Saved in instrument state.

50

–200

Prior to A.02.00

A.02.00

RRC Weighting (All Offsets)

This key turns on or off RRC weighting for the power measurement for all offsets. If RRC weighting is turned on, but you wish to exclude RRC weighting for a particular offset, set its filter alpha to 0.

Key Path

Mode

Measurement

Remote Command

Trace/Detector,ACP,Offsets

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:FILTer:

RRC:STATe OFF|ON|0|1

:CALCulate:<meas>:TRACe[1]|2|3|4:ACPower:OFFSet:FILTer:

RRC:STATe?

462

Common Measurement Functions 2

Example

Preset

State Saved

Initial S/W Revision

Modified at S/W Revision

CALC:VECT:TRAC1:ACP:OFFS:FILT:RRC:STAT ON

CALC:VECT:TRAC1:ACP:OFFS:FILT:RRC:STAT?

0

Saved in instrument state.

Prior to A.02.00

A.02.00

OBW Setup (Occupied Bandwidth)

The occupied bandwidth (OBW) function finds and displays the band of frequencies that contain a specified percentage of the total power within the measurement span.

The OBW key accesses a menu of functions that allow you to define and turn on the OBW function on the selected trace.

An OBW measurement may be defined for each trace, although it will only be active on frequency-domain trace data. The band defined by the OBW measurement is shown as a blue bar overlaying the trace display. To see tabular data showing the frequencies of the band limits, the total power, etc. you may assign the OBW Summary (Trace n) to a different trace. For example, you can assign Spectrum data to trace 3, turn on OBW on trace 3, and assign the OBW Summary (Trace 3) to trace 4, as shown below.

463

Common Measurement Functions 2

The summary data may be retrieved programmatically using FETCh? or the

CALCulate:<meas>:DATA:TABLe commands. See the Data Queries section under Common Functions for more details.

Key Path

Mode

Readback Text

Initial S/W Revision

Modified at S/W Revision

Trace/Detector

VSA, LTE, LTETDD, IDEN

[On|Off, <num>%]

Prior to A.02.00

A.02.00

OBW Summary for Trace 1

Select trace for assignment of OBW Data.

See also: “OBW Setup (Occupied Bandwidth)” on page 463

Key Path

Trace/Detector, Data, OBW, Channel n

464

Common Measurement Functions 2

Mode

Initial S/W Revision

Modified at S/W Revision

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

OBW Summary for Trace 2

Select trace for assignment of OBW Data

See also:

“OBW Setup (Occupied Bandwidth)” on page 463 .

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data, OBW, Channel n

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

OBW Summary for Trace 3

Select trace for assignment of OBW Data

See also:

“OBW Setup (Occupied Bandwidth)” on page 463 .

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data, OBW, Channel n

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

OBW Summary for Trace 4

Select trace for assignment of OBW Data

See also:

“OBW Setup (Occupied Bandwidth)” on page 463 .

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data, OBW, Channel n

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

OBW Summary for Trace 5

Select trace for assignment of OBW Data

See also:

“OBW Setup (Occupied Bandwidth)” on page 463 .

Key Path

Trace/Detector, Data, OBW, Channel n

465

Common Measurement Functions 2

Mode VSA

OBW Summary for Trace 6

Select trace for assignment of OBW Data

See also: “OBW Setup (Occupied Bandwidth)” on page 463

.

Key Path

Mode

Trace/Detector, Data, OBW, Channel n

VSA

OBW Power

The OBW Power key is used to specify the percentage of power used to determine the occupied BW, and to turn the OBW function on or off for the selected trace.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Couplings

Preset

State Saved

Min

Max

Initial S/W Revision

Trace/Detector, OBW

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:PERCent <real>

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:PERCent?

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:STATe

OFF|ON|0|1

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:STATe?

CALC:VECT:TRAC1:OBW:PERC 99

CALC:VECT:TRAC1:OBW:PERC?

CALC:VECT:TRAC1:OBW:STAT ON

CALC:VECT:TRAC1:OBW:STAT?

Parameter is interpreted as a percent, e.g., if you want the OBW to be 95% send 95, not 0.95

None

Controls the presence or absence of data in the OBW Summary table for the selected trace

99.0

0

Saved in instrument state.

0

100

Prior to A.02.00

466

Common Measurement Functions 2

Modified at S/W Revision A.02.00

OBW Centroid > CF

This softkey is used to copy the centroid of the occupied bandwidth to the Center Frequency. It only works if the currently selected trace has data compatible with the OBW function, and OBW is turned on.

This is a front-panel function only.

You can read the OBW centroid using the following SCPI-only query and use the result to set the center frequency.

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:CENTroid?

CALC:VECT:TRAC1:OBW:CENT?

Query only. Returns NaN (9.91E+37) if the OBW function is not active for the selected trace, or is not supported for the trace data assigned to the selected trace.

Prior to A.02.00

A.02.00

BW Limit

This turns on or off limit testing for the Occupied BW test for the selected trace, and allows you to define the limit. Test pass or fail status appears in the OBW Summary table associated with the trace.

Key Path

Mode

Measurement

Remote Command

Example

Trace/Detector, OBW

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:LIMit:FBLimit

<freq>

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:LIMit:FBLimit?

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:LIMit[:TEST]

OFF|ON|0|1

:CALCulate:<meas>:TRACe[1]|2|3|4:OBWidth:LIMit[:TEST]?

CALC:VECT:TRAC1:OBW:LIMIT:FBL 10 MHZ

CALC:VECT:TRAC1:OBW:LIMIT:FBL?

CALC:VECT:TRAC1:OBW:LIMIT:TEST ON

CALC:VECT:TRAC1:OBW:LIMIT:TEST?

467

Common Measurement Functions 2

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

1000000

0

Saved in instrument state.

1 Hz

9.9e37 (Infinity) Hz

Prior to A.02.00

A.02.00

Register

Accesses a menu that allows you to select registers for assignment of trace data.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Data 1

Select register 1 for assignment of trace data.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data, Register

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Data 2

Selects register 1 for assignment of trace data.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data, Register

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Data 3

Selects register 1 for assignment of trace data.

Key Path

Trace/Detector, Data, Register

468

Common Measurement Functions 2

Mode

Initial S/W Revision

Modified at S/W Revision

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Data 4

Selects register 1 for assignment of trace data.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data, Register

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Data 5

Selects register 5 for assignment of trace data.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data, Register

VSA

Prior to A.02.00

A.02.00

Data 6

Selects register 6 for assignment of trace data.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trace/Detector, Data, Register

VSA

Prior to A.02.00

A.02.00

Trace Indicator Info

This softkey allows you to get more information about why a trace indicator is showing. A trace indicator appears in the upper right corner of a trace display to announce exceptional conditions. When such an indicator is showing on the selected trace, pressing this key causes more information about the condition to appear in the message area. This is a front-panel only function. The SCPI commands for querying the Trace Indicator and the Trace Indicator Info for a particular trace are:

CALC:<meas>:DATA[1]|2|3|4:HEAD:STR? "TrcLedStr"

CALC:<meas>:DATA[1]|2|3|4:HEAD:STR? "TrcLedReason"

Key Path

Trace/Detector

469

Common Measurement Functions 2

Mode

Initial S/W Revision

Modified at S/W Revision

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Trigger

Triggering is used to determine when a measurement should start taking data. There are several available trigger sources. For each trigger source, there are associated setup parameters. Typically, a trigger event is generated when a signal (or a characteristic of the signal) crosses a defined trigger level (or threshold) on a rising or falling slope. The measurement begins at a specified time delay from the trigger point. The delay may be negative, allowing pre-trigger data to be taken. Each trigger source has associated its own trigger level, slope, and delay settings.

Trigger Holdoff - Some form of trigger holdoff is available for most trigger types. Hold off can be

defined in different ways, with possible variations depending on trigger slope setting.

Normal: This is the holdoff type that scopes typically use. After a trigger event, for the duration of the holdoff time, no additional trigger events are recognized.

Below Level: If the trigger slope is positive, a trigger event is generated only if the signal characteristic of interest crosses the trigger threshold (with positive slope) after having been below the threshold for at least the holdoff time. For negative slope, the trigger event is generated if the signal characteristic crosses the threshold (with negative slope) and then remains below the threshold for at least the holdoff time. In either case, the trigger event is associated with the time the level was crossed.

Above Level: If the trigger slope is positive, a trigger event is generated only if the signal characteristic of interest crosses the trigger threshold (with positive slope) and then remains above the threshold for at least the holdoff time. For negative slope, the trigger event is generated if the signal characteristic crosses the threshold (with negative slope) after having been above the threshold for at least the holdoff time. In either case, the trigger event is associated with the time the level was crossed.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Front Panel

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Trig Reference Line

The trigger reference line appears (if enabled) when the trigger source is related to the measured signal.

It shows the trigger level relative to the signal. This control allows you to show or hide the trigger reference line.

The trigger reference line, only appears on appropriately formatted time traces. For example, if Video (IF

Envelope) trigger is selected, the trigger level line would appear on Main Time, Inst Main Time, or Raw

Main Time traces that are formatted as Log Mag or Linear Mag.

Key Path

Trigger

470

Common Measurement Functions 2

Mode

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

VSA, LTE, LTETDD, IDEN

:TRIGger[:SEQuence]:RLINe OFF|ON|0|1

:TRIGger[:SEQuence]:RLINe?

TRIG:RLIN ON

TRIG:RLIN?

1

Saved in instrument state.

Show | Hide

Prior to A.02.00

A.02.00

Hardware Trigger

When the Data Source is Inputs, this trigger menu appears. The menu gives you a choice of trigger sources. Once you select a trigger source, you can branch to the setup parameters for that source.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Trigger

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:TRIGger:<meas>[:SEQuence]:SOURce

IMMediate|VIDeo|EXTernal1

:TRIGger:<meas>[:SEQuence]:SOURce?

TRIG:VECT:SOUR IMM

TRIG:VECT:SOUR?

Video triggering is also known by some as IF Envelope or IF Magnitude triggering.

IMM

Saved in instrument state.

Free Run | Video (IF Envelope) | External 1

Prior to A.02.00

A.02.00

Free Run

Free Run triggering, means each measurement scan starts as soon as possible, without regard to any

471

Common Measurement Functions 2

signal characteristics or external triggering signal.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trigger

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Video (IF Envelope)

Pressing this key, when it is not selected, selects Video (IF Envelope) triggering. The trigger condition is met when the magnitude of the signal you are measuring crosses the defined trigger level while satisfying the slope and holdoff conditions. (Specifically, the source for the trigger calculation is the IF signal, filtered only by the brickwall filter that defines the information bandwidth of the signal, Signal energy outside the information bandwidth does not affect the triggering.)

NOTE

This is called Video triggering due to its similarity with swept analyzer zero span measurements being triggered on the video signal. However, in this case there is no video signal. Since the trigger condition applies to the full IF signal, this is also called IF envelope triggering.

If Video triggering is already selected then pressing this softkey accesses the video trigger setup functions, and changes the active function to Video Trigger Level.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trigger

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Trigger Level

Sets a level (in volts) that the magnitude of the IF signal must cross (with the correct slope) in order to generate a trigger. (Holdoff conditions must also be met.)

Key Path

Mode

Remote Command

Example

Notes

Preset

Trigger,Video

VSA, LTE, LTETDD, IDEN

:TRIGger[:SEQuence]:VIDeo:LEVel <voltage>

:TRIGger[:SEQuence]:VIDeo:LEVel?

TRIG:VID:LEV 10 MV

TRIG:VID:LEV?

:TRIGger[:SEQuence]:IF:LEVel <voltage> may be used as an alias

10 mV

472

Common Measurement Functions 2

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Saved in instrument state.

0

9.9E+37

Prior to A.02.00

A.02.00

Trig Slope

Controls the trigger polarity. Positive means the trigger occurs when the rising magnitude crosses the trigger level.

Negative means the trigger occurs when the falling magnitude crosses the trigger level.

Key Path

Mode

Remote Command

Example

Notes

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Trigger,Video

VSA, LTE, LTETDD, IDEN

:TRIGger[:SEQuence]:VIDeo:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:VIDeo:SLOPe?

TRIG:VID:SLOP POS

TRIG:VID:SLOP?

:TRIGger[:SEQuence]:IF:SLOPe POSitive|NEGative may also be used

POS

Saved in instrument state.

Pos | Neg

Prior to A.02.00

A.02.00

Trig Delay

Controls the time delay from the trigger point to the actual start of the measurement data. This can be negative to get pre-trigger data.

Key Path

Mode

Remote Command

Trigger,Video

VSA, LTE, LTETDD, IDEN

:TRIGger[:SEQuence]:VIDeo:DELay <time>

:TRIGger[:SEQuence]:VIDeo:DELay?

:TRIGger[:SEQuence]:VIDeo:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:VIDeo:DELay:STATe?

473

Common Measurement Functions 2

Example

Notes

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

.

Trig Holdoff

Sets the trigger holdoff time.

Key Path

Mode

Remote Command

TRIG:VID:DEL 10 MS

TRIG:VID:DEL?

TRIG:VID:DEL:STAT ON

TRIG:VID:DEL:STAT?

:TRIGger[:SEQuence]:IF:DELay <time> may be used as an alias

:TRIGger[:SEQuence]:IF:DELay:STATe may also be used as an alias

0

OFF

Saved in instrument state.

–9.9E+37

9.9E+37

Prior to A.02.00

A.02.00

Example

Notes

Preset

State Saved

Min

Trigger,Video

VSA, LTE, LTETDD, IDEN

:TRIGger[:SEQuence]:VIDeo:HOLDoff <time>

:TRIGger[:SEQuence]:VIDeo:HOLDoff?

:TRIGger[:SEQuence]:VIDeo:HOLDoff:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:VIDeo:HOLDoff:STATe?

TRIG:VID:HOLD 1 US

TRIG:VID:HOLD?

TRIG:VID:HOLD:STAT ON

TRIG:VID:HOLD:STAT?

:TRIGger[:SEQuence]:IF:HOLDoff may be used as an alias

:TRIGger[:SEQuence]:IF:HOLDoff:STATe may be used as an alias

0

OFF

Saved in instrument state.

–9.9E+37

474

Common Measurement Functions 2

Max

Initial S/W Revision

Modified at S/W Revision

9.9E+37

Prior to A.02.00

A.02.00

Holdoff Type

Sets the trigger holdoff type.

Some form of trigger holdoff is available for most trigger types. Hold off can be defined in different ways, with possible variations depending on trigger slope setting.

Below Level: If the trigger slope is positive, a trigger event is generated only if the signal characteristic of interest crosses the trigger threshold (with positive slope) after having been below the threshold for at least the holdoff time. For negative slope, the trigger event is generated if the signal characteristic crosses the threshold

(with negative slope) and then remains below the threshold for at least the holdoff time. In either case, the trigger event is associated with the time the level was crossed.

Above Level: If the trigger slope is positive, a trigger event is generated only if the signal characteristic of interest crosses the trigger threshold (with positive slope) and then remains above the threshold for at least the holdoff time. For negative slope, the trigger event is generated if the signal characteristic crosses the threshold

(with negative slope) after having been above the threshold for at least the holdoff time. In either case, the trigger event is associated with the time the level was crossed.

Key Path

Mode

Remote Command

Example

Notes

Preset

State Saved

Range

Initial S/W Revision

Modified at S/W Revision

Trigger,Video

VSA, LTE, LTETDD, IDEN

:TRIGger[:SEQuence]:VIDeo:HOLDoff:TYPE BELow|ABOVe

:TRIGger[:SEQuence]:VIDeo:HOLDoff:TYPE?

TRIG:VID:HOLD:TYPE BEL

TRIG:VID:HOLD:TYPE?

:TRIGger[:SEQuence]:IF:HOLDoff:TYPE can be used as an alias

BEL

Saved in instrument state.

Below Level | Above Level

Prior to A.02.00

A.02.00

External 1

Pressing this key, when it is not selected, selects the signal on the Trigger 1 input as the trigger signal.

The trigger condition is met when the level of the external trigger signal crosses the defined trigger level while satisfying the slope and holdoff conditions.

Note that currently, the VSA based measurements do not support External 2 triggering.

If External 1 triggering is already selected then pressing this softkey accesses the external 1 trigger setup

475

Common Measurement Functions 2

functions, and changes the active function to Ext 1 Trigger Level.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Trigger

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Trigger Level

Sets a level (in volts) that the Trigger signal must cross (with the correct slope) in order to generate a trigger.

(Holdoff conditions must also be met.)

Key Path

Mode

Remote Command

Example

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Trigger, External 1

VSA, LTE, LTETDD, IDEN

:TRIGger[:SEQuence]:EXTernal1:LEVel <voltage>

:TRIGger[:SEQuence]:EXTernal1:LEVel?

TRIG:EXT1:LEV 10 MV

TRIG:EXT1:LEV?

1 V

Saved in instrument state.

–9.9E+37

9.9E+37

Prior to A.02.00

A.02.00

Trig Slope

Controls the trigger polarity. Positive means the trigger occurs on a rising edge. Negative means the trigger occurs on a falling edge.

Key Path

Mode

Remote Command

Example

Preset

State Saved

Trigger, External 1

VSA, LTE, LTETDD, IDEN

:TRIGger[:SEQuence]:EXTernal1:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:EXTernal1:SLOPe?

TRIG:EXT1:SLOP POS

TRIG:EXT1:SLOP?

POS

Saved in instrument state.

476

Common Measurement Functions 2

Range

Initial S/W Revision

Modified at S/W Revision

Pos | Neg

Prior to A.02.00

A.02.00

Trig Delay

Controls the time delay from the trigger point to the actual start of the measurement data. This can be negative to get pre-trigger data.

Key Path

Mode

Remote Command

Example

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

Trig Holdoff

Sets the trigger holdoff time.

Trigger, External 1

VSA, LTE, LTETDD, IDEN

:TRIGger[:SEQuence]:EXTernal1:DELay <time>

:TRIGger[:SEQuence]:EXTernal1:DELay?

:TRIGger[:SEQuence]:EXTernal1:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:EXTernal1:DELay:STATe?

0

0

TRIG:EXT1:DEL 10 MS

TRIG:EXT1:DEL?

TRIG:EXT1:DEL:STAT ON

TRIG:EXT1:DEL:STAT?

Saved in instrument state.

–9.9E+37

9.9E+37

Prior to A.02.00

A.02.00

Key Path

Mode

Remote Command

Trigger, External 1

VSA, LTE, LTETDD, IDEN

:TRIGger[:SEQuence]:EXTernal1:HOLDoff <time>

:TRIGger[:SEQuence]:EXTernal1:HOLDoff?

:TRIGger[:SEQuence]:EXTernal1:HOLDoff:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:EXTernal1:HOLDoff:STATe?

477

Common Measurement Functions 2

Example

Preset

State Saved

Min

Max

Initial S/W Revision

Modified at S/W Revision

0

0

TRIG:EXT1:HOLD 1 US

TRIG:EXT1:HOLD?

TRIG:EXT1:HOLD:STAT ON

TRIG:EXT1:HOLD:STAT?

Saved in instrument state.

–9.9E+37

9.9E+37

Prior to A.02.00

A.02.00

Holdoff Type

Sets the trigger holdoff type. See

Some form of trigger holdoff is available for most trigger types. Hold off can be defined in different ways, with possible variations depending on trigger slope setting.

Below Level: If the trigger slope is positive, a trigger event is generated only if the signal characteristic of interest crosses the trigger threshold (with positive slope) after having been below the threshold for at least the holdoff time. For negative slope, the trigger event is generated if the signal characteristic crosses the threshold (with negative slope) and then remains below the threshold for at least the holdoff time. In either case, the trigger event is associated with the time the level was crossed.

Above Level: If the trigger slope is positive, a trigger event is generated only if the signal characteristic of interest crosses the trigger threshold (with positive slope) and then remains above the threshold for at least the holdoff time.

For negative slope, the trigger event is generated if the signal characteristic crosses the threshold (with negative slope) after having been above the threshold for at least the holdoff time. In either case, the trigger event is associated with the time the level was crossed.

Key Path

Mode

Remote Command

Example

Preset

State Saved

Range

Initial S/W Revision

Trigger, External 1

VSA, LTE, LTETDD, IDEN

:TRIGger[:SEQuence]:EXTernal1:HOLDoff:TYPE BELow|ABOVe

:TRIGger[:SEQuence]:EXTernal1:HOLDoff:TYPE?

TRIG:EXT1:HOLD:TYPE BEL

TRIG:EXT1:HOLD:TYPE?

BEL

Saved in instrument state.

Below Level | Above Level

Prior to A.02.00

478

Common Measurement Functions 2

Modified at S/W Revision A.02.00

View/Display

The View/Display menu contains the Display branch key, which allows you to set many display properties. Many View Preset softkeysappear under this menu. These set up measurement-specific views, which are described in individual measurements. A view in this application is simply a preset; i.e., a choice of layout, trace data assignment, and trace formatting and scaling. After a view preset is performed, the resulting arrangement can then be changed by any available trace manipulation functions or by changing the layout. All measurements have a default view that is used when they are first started, and the first listed preset view will restore that arrangement without otherwise affecting the measurement.

This menu contains keys that allow control over the way data is displayed. The Layout key is described here. Other keys specific to measurements will be described in their own descriptions.

Key Path

Mode

Initial S/W Revision

Modified at S/W Revision

Front Panel

VSA, LTE, LTETDD, IDEN

Prior to A.02.00

A.02.00

Layout

This key allows you to choose the number and position of windows on the screen. Each window contains one trace. The selected trace is always visible and its window outlined in green. The Window zoom key toggles between multiple windows and a single window mode without changing the setting for Layout.

Single layout has one window.

Stack 2 layout has two windows, one on top of the other, that display either traces 1 (top) and 2 (bottom) or traces 3 and 4. The pair that is showing always includes the selected trace.

Stack 3 layout has three windows that display, top to bottom, traces 1, 2, 3 or traces 2, 3, 4.

Grid 2x2 layout has 4 windows, arranged 2x2.They display (in order top to bottom, left to right) traces 1,

479

Common Measurement Functions 2

2, 3, and 4.

Grid 2x2 layout with Trace 2 selected

There are two other layouts that are available for iDEN Power, iDEN Demod and MOTOTalk measurements, since these allow 6 traces.

Grid 2x3 layout has 2 rows of 3 windows that display all 6 traces in order, top to bottom, them left to right.

Grid 3x2 layout has 3 rows of 2 windows that display all 6 traces in order, top to bottom, them left to right.

480

Common Measurement Functions 2

iDEN Demod, iDEN Power, and MOTOtalk

Layout

Layout

Single

Stack 2

Stack 3

Grid 2x2

Grid 2x3

Grid 3x2

Key Path

Mode

Measurement

Remote Command

Example

Couplings

Preset

State Saved

Range

View/Display

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:DISPlay:<meas>:WINDow:FORMat SINGle|TWO|TRI|QUAD

:DISPlay:<meas>:WINDow:FORMat?

For iDEN Power, iDEN Demod and MotoTalk measurements:

:DISPlay:<meas>:WINDow:FORMat

SINGle|TWO|TRI|QUAD|GR2X3|GR3X2

:DISPlay:<meas>:WINDow:FORMat?

DISP:VECT:WIND:FORM TWO

DISP:IPOW:WIND:FORM GR2X3

DISP:VECT:WIND:FORM?

If the window is currently zoomed, selecting a layout (even the current one) will switch it to tiled mode.

TWO|QUAD|QUAD|QUAD|QUAD|QUAD|QUAD|QUAD|GR2X3|TRI

Saved in instrument state.

Single | Stack 2 | Stack 3 | Grid 2x2 | Grid 2x2 | Grid 2x3 | Stack 3

481

Common Measurement Functions 2

Initial S/W Revision

Modified at S/W Revision

Prior to A.02.00

A.02.00

Remote SCPI Commands and Data Queries

Remote SCPI Results described in this section include:

“:READ and :FETCh Commands” on page 482

“:CALCulate:DATA” on page 486

“:CALCulate:DATA:RAW” on page 488

“:CALCulate:DATA:RAW:COMPlex” on page 488

“:CALCulate:DATA:POINts commands” on page 488

“:CALCulate:DATA:TABL commands” on page 489

“:CALCulate:DATA:HEADer commands” on page 493“IQ Data Transfers ” on page 495

VSA based Measurements produce a rich variety of results which may be displayed in any of 4 traces. A result may consist of an array of X,Y trace data that is typically shown as a graph, or scalar results that are displayed as a table. The Symbol/Error result that is part of many demodulation measurements actually displays both a trace table (the error statistics) and trace data (the symbol information, which is not graphed but listed). The CALC:<meas>:DATA<n> commands allow you to retrieve any trace data or trace table. This family of commands also allow you to get information about the names of data results available and the units associated with them, as well as names and results of meta-data associated with traces.

Selected results are available via the FETCh and READ SCPI interfaces. These commands refer to data results by arbitrary index number rather than by trace number.

:READ and :FETCh Commands

The SCPI MEASure, READ, and FETCh are typically offered by applications with focus on manufacturing test, where a fixed set of desired results is known in advance and seldom changes. The

VSA based measurements are many, due to a focus on development. Thus, for most VSA based measurements there is no standard configuration that will yield a useful measurement 90% of the time.

Thus, the MEASure function will not be offered for most measurements in the VSA Application.

However, READ and FETCh may be implemented for select results. Note that these results will also still be available using the CALC:<meas>:DATA:TABLe family of commands.

ACP and OBW are available in all VSA based measurements. To retrieve the ACP or OBW data, the

function must be enabled on a frequency-domain trace and the associated summary data table must be assigned to another trace. Note however, the index n in the following commands is not trace number, but an index picked out of the tables shown below.

:FETCh:<meas>[n]?

:READ:<meas>[n]?

482

Common Measurement Functions 2

The results available for various values of n are shown below:

Condition N

Mode = VSA |

LTE | IDEN

Not specified or n=1

Mode = VSA |

LTE | IDEN

2 – 50

Results Returned

Reserved for selected results of VSA measurements.

If not used for a particular measurement, no result is returned and error –114

Header suffix out of range is generated

Reserved for selected results of VSA measurements.

If not used for a particular measurement, no result is returned and error –114

Header suffix out of range is generated

483

Common Measurement Functions 2

Condition

Mode = VSA |

LTE | IDEN,

ACP on trace 1

Mode = VSA |

LTE | IDEN,

ACP on trace 2

N

51

52

Results Returned

ACP Summary for trace 1

Returns 28 comma-separated scalar results, corresponding to the swept ACP results where possible; n/a elsewhere:

Returns 28 comma-separated scalar results, in the following order.

1. 0.0

2. Total carrier power (dBm) (same as item 4, because only 1 carrier supported)

3. 0.0

4. Reference carrier power (dBm)

5. Lower offset A - relative power (dB)

6. Lower offset A - absolute power (dBm)

7. Upper offset A - relative power (dB)

8. Upper offset A - absolute power (dBm)

9. Lower offset B - relative power (dB)

10. Lower offset B - absolute power (dBm)

11. Upper offset B - relative power (dB)

12. Upper offset B - absolute power (dBm)

...

21. Lower offset E - relative power (dB)

22. Lower offset E - absolute power (dBm)

23. Upper offset E - relative power (dB)

24. Upper offset E - absolute power (dBm)

25. n/a

26. n/a

27. n/a

28. n/a

29. Overall ACP test result summary (0 indicates at least 1 failure, 1 indicates all passed)

If any result is not available, NaN (9.91 E 37) is returned. This can happen if ACP is turned off (all results unavailable), or when an offset is entirely off-screen. In the case where it is partially off-screen, the measured result is returned, even though its validity is questionable.

ACP Summary for trace 2 see list for trace 1 summary

484

Condition N

Mode = VSA |

LTE | IDEN,

ACP on trace 3

Mode = VSA |

LTE | IDEN,

ACP on trace 4

Mode = VSA |

LTE | IDEN,

ACP on trace 5

Mode = VSA |

LTE | IDEN,

ACP on trace 6

53

54

55

56

Mode = VSA |

LTE | IDEN,

OBW on trace

1

57–60

61

Mode = VSA |

LTE | IDEN,

OBW on trace

2

Mode = VSA |

LTE | IDEN,

OBW on trace

3

Mode = VSA |

LTE | IDEN,

OBW on trace

4

62

63

64

Common Measurement Functions 2

Results Returned

ACP Summary for trace 3 see list for trace 1 summary

ACP Summary for trace 4 see list for trace 1 summary

ACP Summary for trace 5 see list for trace 1 summary

ACP Summary for trace 6 see list for trace 1 summary no result returned; error –114, Header suffix out of range generated

OBW Summary for trace 1

Returns 9 comma-separated scalar results, corresponding exactly to the items in the

OBW Summary trace:

1. OBW (Hz)

2. Pwr (dBm)

3. Total Pwr (dBm)

4. Pwr Ratio (no unit, E.g. 0.99)

5. OBW upper freq (Hz)

6. OBW lower freq (Hz)

7. Centroid freq (Hz)

8. Offset freq (Hz)

9. OBW Test Result (0 for fail, 1 for pass)

If the results are not available, NaN (9.91 E 37) is returned.

OBW Summary for trace 2 see list for trace 1 summary

OBW Summary for trace 3 see list for trace 1 summary

OBW Summary for trace 4 see list for trace 1 summary

485

Common Measurement Functions 2

Condition N

Mode = VSA |

LTE | IDEN,

OBW on trace

5

Mode = VSA |

LTE | IDEN,

OBW on trace

6

65

66

Results Returned

OBW Summary for trace 5 see list for trace 1 summary

OBW Summary for trace 6 see list for trace 1 summary

Key Path

Mode

Remote Command Only

LTE, LTETDD, IDEN, VSA

:CALCulate:DATA

Once measurement data result is assigned to a trace, the data can be retrieved by using one of the following commands (where <n> is the trace number and <meas> is the current VSA based measurement).

:CALC:<meas>:DATA<n>?

:CALC:<meas>:DATA<n>:RAW?

The first form of the command retrieves the data as formatted on the display. For example, if (in a vector measurement) you have the Spectrum result in LogMag format on trace 1, then

:CALC:VECT:DATA1?

will return an array of spectrum amplitude (Y data) in units of dBm, and

:CALC:VECT:DATA1:RAW?

will return the Y data in its underlying units of Volts (peak) squared.

(To get data from displayed tables, see CALCulate:DATA:TABLe commands below.)

The CALC:<meas>:DATA commands get data from traces. There are many results available from a

VSA based measurement, and only 4 traces in which to view them. View Preset commands are one way of displaying frequently-used results in standard trace locations. Or you may assign any measurement result to any trace using the softkeys under Trace/Detector, Data. The SCPI command for doing this is:

:DISP:<meas>:TRAC<n>:FEED "<data_name>"

For example, if (in a vector measurement) you wish to view the CCDF result in trace 4, you send:

:DISP:VECT:TRAC4:FEED "CCDF1"

(If the measurement has not run yet, use INIT:IMM to run it.) Then the CCDF data may be retrieved using

CALC:VECT:DATA4?

or

486

Common Measurement Functions 2

CALC:VECT:DATA4:RAW?

See the Data command (under Trace/Detector) and the View Preset commands (under View/Display) in this document as well as the PDs for each VSA based measurement for more details on assigning data to traces.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

SCPI only

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:DATA[1]|2|3|4? [Y|X|XY[,OFF|ON|0|1]]

CALC:VECT:DATA1?

CALC:VECT:DATA1? Y,ON

CALC:VECT:DATA1? X

CALC:VECT:DATA1? XY

Query only. This retrieves the data in the designated trace as displayed.

E.g., if Trace 1 is assigned Spectrum data and formatted as LogMag, then

:CALC:VECT:DATA1? will return the Y data in dBm. If the X axis is scaled to show only a portion of the trace data, only the data shown will be returned.

The numeric format of the returned data is controlled by

FORMat[:TRACe][:DATA] command

The optional parameters control what data is returned.

:CALC:VECT:DATA1? Y is the same as :CALC:VECT:DATA1? with no parameter. It returns an array of Y values.

:CALC:VECT:DATA1? X returns an array of X values that correspond to the

Y values above.

:CALC:VECT:DATA1? XY returns interleaved X and Y data. I.e.:

<x1><y1><x2><y2>…

Normally, this command only returns the data between the current X scale limits. If the optional ",OFF" or ",0" switch is included at the end of the command, then all data is returned (regardless of X scaling or the state of All

Frequency Points).

Note: the X and Y parameters in this command refer to the display's horizontal and vertical axes. Normally the X axis is the independent variable, but if the display format is Constellation or IQ, then

CALC:<meas>:DATA<n>? [Y] returns the imaginary part of the data, and

CALC:<meas>:DATA<n>? X returns the real part of the data. If you want the values of the independent variable, change to a non-vector format (such as

Log Mag) and use CALC:<meas>:DATA<n>? X

Prior to A.02.00

A.02.00

487

Common Measurement Functions 2

:CALCulate:DATA:RAW

Retrieves trace data in its underlying units, before the formatting calculation that converts it to displayed units. Underlying units are typically Volts peak (for signal results) or Volts peak squared (for power results). All data points are returned, whether or not they are displayed.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

SCPI only

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:DATA[1]|2|3|4:RAW?

CALC:VECT:DATA1:RAW?

Query only. This retrieves the unformatted Y data in the designated trace. If Y data is complex, it is returned as <y_real1><y_imag1><y_real2><y_imag2> etc.

Prior to A.02.00

A.02.00

:CALCulate:DATA:RAW:COMPlex

This command is used to determine if the data retrieved by CALC:<meas>:DATA:RAW<n>? is complex.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

SCPI only

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:DATA[1]|2|3|4:RAW:COMPlex?

CALC:VECT:DATA1:RAW:COMP?

Query only. Returns 1 if the trace data is complex, 0 if it is real.

Prior to A.02.00

A.02.00

:CALCulate:DATA:POINts commands

This query returns the number of points that will be returned by

CALCulate:<meas>:DATA<n>?

X axis scaling and whether All Frequency Points is on or off can affect this number.

NOTE

For the CALCulate:<meas>:DATA<n>? XY command there are 2 numbers

488

Common Measurement Functions 2

returned per data point.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

SCPI only

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:DATA[1]|2|3|4:POINts? [OFF|ON|0|1]

CALC:VECT:DATA1:POINts?

Query only.

Use the optional "OFF|0" parameter to determine the number of points that will be returned by the the optional command form:

:CALCulate:<meas>:DATA<n>? Y|X|XY,OFF|0

Note that this is points, not array size. If the XY parameter is included, there are 2 numbers returned per point.

(ON or 0, which means use the X-scaled version, is the default and the result is the same as if the parameter is omitted).

Prior to A.02.00

A.02.00

Key Path

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

SCPI only

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:DATA[1]|2|3|4:RAW:POINts?

CALC:VECT:DATA1:RAW:POINts?

Query only.

Prior to A.02.00

A.02.00

:CALCulate:DATA:TABL commands

Query Table Data as Number

The following query gets data from a table shown in the designated trace. Tables shown on the display typically have the name of a parameter followed by its measured value

Key Path

Mode

SCPI only

VSA, LTE, LTETDD, IDEN

489

Common Measurement Functions 2

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:DATA[1]|2|3|4:TABLe[:NUMBer]?

[<string>]

CALC:DDEM:DATA2:TABL? "Obw"

Query only. If sent without a string specifier, this returns the entire table for the designated trace.If sent with a string specifier, returns a specific table entry in the designated trace. The string specifier must be delimited by single or double quotes. A list of valid strings can be obtained using

CALC:<meas>:DATA:TABL:NAM? If an invalid string is sent, an error is generated. The returned results are in numeric format, under control of the

FORMat[:TRACe][:DATA] command. For table data that is non-numeric,

NaN is returned. To get the value of these data, use the

CALC:<meas>:DATA2:TABL:STR? command

Prior to A.02.00

A.02.00

Query Table Data as String

Some tables have string data. The above Trace Table Data query cannot return it, and sends NaN in its place. Here is a form of Trace Table Data query that can return string data from tables.

Key Path

Mode

Measurement

Remote Command

Example

Notes

SCPI only

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:DATA[1]|2|3|4:TABLe:STRing?

[<string>]

CALC:DDEM:DATA2:TABL:STR? "Obw"

Query only. If sent without a string specifier, this returns the entire table for the designated trace, in comma-separated format. If sent with a string specifier, returns a specific table entry in the designated trace. The string specifier must be delimited by single or double quotes. A list of valid strings can be obtained using CALC:<meas>:DATA:TABL:NAM? If an invalid string is sent, an error is generated.

Prior to A.02.00

A.02.00

Initial S/W Revision

Modified at S/W Revision

Query Table Names

The following query returns a comma-separated list of names of the table data entries for the designated trace. Each of names may be used (surrounded by quotes or double quotes) as a parameter in the Trace

Table Data commands. The names appear in the same order as the corresponding data values returned by

490

Common Measurement Functions 2

the CALC:<meas>:DATA<n>:TABL[:NUMB|STR]? query.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

SCPI only

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:DATA[1]|2|3|4:TABLe:NAMes?

CALC:VECT:DATA1:TABL:NAM?

Query only. This retrieves the names of the table entries for the designated trace. Each of these names may be used in the CALC:<meas>:DATA:TABL?

'<name>' command to access a single table entry.

Prior to A.02.00

A.02.00

Query Table Units

The following query returns a comma-separated list of all the units for the table data entries for the designated trace. If a data result is unitless, an empty string appears in the list for that result. The units appear in the same order as the corresponding data values returned by the

CALC:<meas>:DATA<n>:TABL[:NUMB|STR]? query.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

SCPI only

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:DATA[1]|2|3|4:TABLe:UNIT?

CALC:VECT:DATA1:TABL:UNIT?

Query only. This retrieves a list of units for table entries for the designated trace. The units are given in the order that the entries are sent from the

:CALC:<meas>:DATA:TABL? command.

Prior to A.02.00

A.02.00

The following table data is available in all measurements when the ACP function is turned on and the associated summary table is shown in a trace:

Result name

Reference Bandwidth

Reference Alpha

Displayed Unit

Hz

Remote Name

RefBw

RefAlpha

Remote Unit

Hz

491

Common Measurement Functions 2

Result name

Reference Power

Offset

BW

Alpha

Lower Pwr

Lower ACPR

Upper Pwr

Upper ACPR

Max ACPR

Displayed Unit

dBm

Hz

Hz dBm dB dBm dB dB

Remote Name Remote Unit

RefPwr

Offset1, Offset2, Offset3,

Offset4, Offset5

Bw1, Bw2, Bw3, Bw4,

Bw5

Alpha1, Alpha2, Alpha3,

Alpha4, Alpha5

Vrms^2

Hz

Hz

LowPwr1, LowPwr2,

LowPwr3, LowPwr4,

LowPwr5

LowRatio1, LowRatio2,

LowRatio3, LowRatio4,

LowRatio5

Vrms^2

HiPwr1, HiPwr2, HiPwr3,

HiPwr4, HiPwr5

Vrms^2

HiRatio1, HiRatio2,

HiRatio3, HiRatio4,

HiRatio5

MaxRatio1, MaxRatio2,

MaxRatio3, MaxRatio4,

MaxRatio5

The following table data is available in all measurements when the OBW function is turned on and the associated summary table is shown in a trace:

Result name

Occupied Bandwidth

Power

Total Power

Power Ratio

Upper Freq

Lower Freq

Centroid Freq

Offset Freq

Displayed Unit

Hz

Hz

Hz

Hz

Hz dBm dBm

%

Remote Name

Obw

Pwr

TotalPwr

PwrRatio

ObwUpper

ObwLower

Centroid

Offset

Hz

Hz

Hz

Hz

Remote Unit

Hz

Vrms^2

Vrms^2

492

Common Measurement Functions 2

:CALCulate:DATA:HEADer commands

Query Header Names

The following query returns a comma-separated list of all the header names associated with the designated trace. Each of the names may be used (surrounded by quotes or double quotes) as a parameter in the other CALC:<meas>:DATA<n>:HEAD queries.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

SCPI only

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:DATA[1]|2|3|4:HEADer:NAMes?

CALC:VECT:DATA1:HEAD:NAM?

Query only. Returns a comma-separated list of header names.

Prior to A.02.00

A.02.00

Query Header Type

This query returns whether the designated header on the designated trace may be queried as a number, or by a string only.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

SCPI only

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:DATA[1]|2|3|4:HEADer:TYPE? <string>

CALC:VECT:DATA1:HEAD:TYPE? 'XDelta'

Query only. This retrieves the type of the named header for the designated trace. The name (delimited by single or double quotes) is one of the names returned by the CALC:<meas>:DATA:HEAD:NAMes?

If a valid header name is passed in, the return value from this query is either

STR or NUMB. NONE is returned if there is no such header.

Prior to A.02.00

A.02.00

Query Header as String

This query gets a header by name from the designated trace and returns its value as a string.

Key Path

SCPI only

493

Common Measurement Functions 2

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:DATA[1]|2|3|4:HEADer:STRing? <string>

CALC:VECT:DATA1:HEAD:STR? 'WindowType'

Query only. This retrieves the named header for the designated trace. The name (delimited by single or double quotes) is one of the names returned by the CALC:<meas>:DATA:HEAD:NAMes? The return value is a string. If the requested header value is a numeric, or if there is no such header, an empty string is returned..

Prior to A.02.00

A.02.00

Query Numeric Header

This query gets a numeric header by name from the designated trace and returns its value in a format determined by the last FORM command.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Initial S/W Revision

Modified at S/W Revision

SCPI only

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

:CALCulate:<meas>:DATA[1]|2|3|4:HEADer[:NUMBer]?

<string>

CALC:VECT:DATA1:HEAD? 'XDelta'

Query only. This retrieves the named header for the designated trace. This form of the HEAD? query is for headers whose type is NUMB (as determined by :CALC:<meas>:DATA:HEAD:TYPE?)

The name parameter (delimited by single or double quotes) is one of the names returned by the CALC:<meas>:DATA:HEAD:NAMes? The format of the return data is determined by the FORMat[:TRACe][:DATA] command.

If used to query a header whose type is STR, or there is no such header, NaN

(9.91e37) is returned

Prior to A.02.00

A.02.00

:CALC:CLIMits:FAIL?

If one or more ACP or OBW limit tests are active, then the CALC:CLIMits:FAIL? command will return the aggregate pass or fail status.

494

Common Measurement Functions 2

IQ Data Transfers

Fast capture/transfer of a large amount of IQ data is supported over SCPI. To do this, first set up the desired measurement range, center frequency, span, triggering, etc. Use a time length that is convenient for setting up the measurement. The time length for the captured data will be set indirectly as shown below.

To perform the capture, a typical SCPI sequence is as follows:

FCAP:LENG <num_samples>

This command sets the length for the next capture in samples. The sample rate is proportional to the current span, and can be determined by a SCPI query, e.g., in the Vector measurement the query:

VECT:SWE:ISR?

returns the input sample rate. Multiply the time length desired for the captured data by this sample rate to get the number of samples to needed.

INIT:FCAP will pause the current measurement and start capturing IQ data using the current setup and trigger conditions. (The instrument front panel display will not change nor show the captured data.)

To read the captured data via SCPI in blocks, set the read block size using the command:

FCAP:BLOC <num_points_per_read_block>

The maximum read block size is typically less than the total fast capture buffer size, and can be determined by the query "FCAP:BLOC? MAX". Now you can repeatedly use the following query to read out successive blocks of data:

FETC:FCAP?

The returned data is formatted according to the most recent :FORMat[:DATA] and :FORMat:BORDer commands. A read pointer that indicates the next sample to be transferred is advanced automatically following each FETC:FCAP? query. This pointer position can be read or manually set via the SCPI commands

FCAP:POIN?

FCAP:POIN <read_pointer_position>

The fast capture data may be read as long as you use only the commands to set read block size and pointer position, or queries that return the state of the current measurement. The capture data is cleared by any command that changes the measurement state or initiates a new measurement, or via SCPI device clear, or :ABORT commands.

Fast capture data word size may be set to either 32 bit or 64 bit via the FCAP:WLEN command. This allows you to trade off precision for total capture length.

NOTE

When the word size is 32 bit, points can only be retrieved on even sample number boundaries, i.e., the pointer and block length should be even numbers. Therefore, when the word size is set to auto, it is recommended that the pointer and block size be only set to even numbers.

495

Common Measurement Functions 2

Fast Capture Length

Sets the length of the SCPI Fast Capture in samples (points)

Query returns the most recent length setting.

Remote Command

Example

Initial S/W Revision

Mode

Notes

Preset

Min

Max

[:SENSe]:FCAPture:LENGth <integer>

[:SENSe]:FCAPture:LENGth?

FCAP:LENG 1000

FCAP:LENG?

A.04.00

VSA

This is affected by the IF path currently used, which may in turn be affected by span. It is also affected by the internal Fast Capture Word Length. The current maximum fast capture length may be found by using the query:

FCAP:LENG? MAX

Changing the Capture Length after initiating a fast capture clears the capture memory in preparation for a new fast capture of a different length.

No Front panel access; SCPI only

1048576 Samples

2

536 870 912 Samples for internal 40 MHz and 140 MHz options with

FCAP:WLEN BIT32

Fast Capture Word Length

Allows choice of internal fast capture word length. Shorter word length allows twice the time length to be captured, at the cost of quantization noise. Note that this does not affect the format of data returned by

FETCh:FCAPture, only the internal representation.

Remote Command

Example

Initial S/W Revision

Mode

Notes

[:SENSe]:FCAPture:WLENgth AUTO|BIT32|BIT64

[:SENSe]:FCAPture:WLENgth?

FCAP:WLEN AUTO

FCAP:WLEN?

A.04.00

VSA

No Front panel access; SCPI only.

Preset AUTO

496

Common Measurement Functions 2

Initiate Fast Capture

Waits for the sweep to trigger and then captures the fast capture data. Sweep is then set to pause. The amount of data captured is controlled by the Fast Capture Length command (FCAP:LENG).

Remote Command:

Example:

Notes:

Initial S/W Revision:

Mode

:INITiate:FCAPture

INIT:FCAP

Returns when the capture is complete.

No Front panel access; SCPI only

This command resets the Fast Capture Pointer to 0

A.04.00

VSA

Fast Capture Block

Sets the block size for the Fast Capture transfer in samples (points). This is the number of points that will be returned from the Capture buffer by the FETC:FCAP? command.

Query returns most recent block size setting.

Remote Command

Example

Initial S/W Revision

Mode

Notes

[:SENSe]:FCAPture:BLOCk <integer>

[:SENSe]:FCAPture:BLOCk?

FCAP:BLOC 100

FCAP:BLOC?

A.04.00

VSA

No Front panel access. SCPI only.

Preset

Min

Max

1024 Samples

0

131072 or Fast Capture Length, whichever is smaller

Fast Capture Pointer

Sets the pointer position for the Fast Capture transfer in samples (points). The pointer is incremented by the block size each time the fetch is performed. Preset value (0) is the first sample in the record. Thus repetitive fetches will result in contiguous data without needing to increment the pointer over SCPI.

Query returns most recent pointer setting.

Remote Command

[:SENSe]:FCAPture:POINter <integer>

[:SENSe]:FCAPture:POINter?

497

Common Measurement Functions 2

Example

Initial S/W Revision

Mode

Notes

FCAP:POIN 100

FCAP:POIN?

A.04.00

VSA

INIT:FCAP or FCAP:ABOR resets the pointer to 0.

No front panel access; SCPI only.

Preset

Min

Max

0 Samples

0

Must be less than the Fast Capture length

Fetch Fast Capture

Transfers the block of data starting at the pointer. The number of samples transferred is set with the block size. The pointer is incremented by the block size after the fetch.

Mode

Remote Command

Example

Notes

Initial S/W Revision

VSA

:FETCh:FCAPture?

FETC:FCAP?

The returned data is formatted according to the most recent :FORMat[:DATA] and :FORMat:BORDer commands.

If the read pointer position plus read block size exceeds the Fast Capture

Length, only the data between the pointer and the end of the fast capture buffer are returned, and error –200 is reported..

If Fetch is attempted before an INIT:FCAP (or if the captured data is cleared by some other operation ,e.g. REC; error –230 is reported and no data is returned.

No front panel access; SCPI only.

A.04.00

Input Sample Rate Query

This SCPI only query returns the complex sample rate in Hz for the current VXA measurement setup conditions. The sample rate can be used to convert between time and number of sample points when using the Fast Capture feature.

Sample rate depends on the settings for FREQ:SPAN and IFPath. You need to set these before making this query. Though the measurement name is specified in the query, you can only query the currently configured measurement. That is, if you have sent CONF:VECT, the query ADEM:SWE:ISR? will generate an error.

Mode VSA

498

Common Measurement Functions 2

Remote Command

Example

Preset

Notes

Initial S/W Revision

[:SENSe]:<meas>:SWEep:ISRate?

VECT:SWE:ISR?

Depends on the licensed IF path

Query returns the complex sample rate in Hz for the current VXA Vector measurement setup conditions.

If the measurement in the query is not the active measurement, error -230 is reported and no data is returned.

This query is SCPI only, no Front Panel softkey.

A.04.00

Parameter Update Enable

When a measurement parameter is changed, the new value is used to update dependent parameters and measurement results. This update process is usually done after every parameter change. This command allows you to postpone this update until after a group of parameters changes, and then allow the update to run once afterwards. This can help decrease setup time for SCPI-controlled measurements.

For example, if your program is setting up a complex LTE measurement, it could save some time by first sending EVM:PUPD:ENAB OFF, then sending the whole group of measurement setup commands.

(Note that if you try to immediately read back a parameter value after setting it, you are not guaranteed to get back true value that will be used in the measurement because dependencies have not been resolved.)

When you are done with the setup, send EVM:PUPD:ENAB:ON. After this, you can read back the parameters' actual values. Starting or continuing a measurement (INITiate:RESTart or

INITiate:IMMediate) will automatically set PUPD:ENAB to ON.

This command should be used carefully. It should not be used before a preset (CONF:<meas>). If you

Parameter Update Enable off, you should make the necessary changes and turn it back on before doing any queries, because no parameter limiting is done nor are any dependencies between parameters resolved while Parameter Update Enable is off.

Key Path

Mode

Measurement

Remote Command

Example

Notes

Preset

State Saved

Initial S/W Revision

SCPI only

VSA, LTE, LTETDD, IDEN

<meas>:=VECTor|ADEMod|DDEMod|W11A|W11B|EVM|IPOWer|IDEMod|

MOTotalk

[:SENSe]:<meas>:PUPDate:ENABle OFF|ON|0|1

[:SENSe]:<meas>:PUPDate:ENABle?

EVM:PUPD:ENAB OFF

INIT:IMM or INIT:REST will set this to ON

1

No

A.03.00

499

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

advertisement

Table of contents