advertisement
NewClassic Balances
MS-S / MS-L Models
C
C
C
C
Table of contents
Table of contents
1
2
3
4
5
6
7
8
Introduction
1.1
Safety Precautions
Overview
3.1
3.2
3.3
3.4
Setting up the Balance
Weighing Made Simple
The Menu
Application "Piece Counting"
Application "Percent Weighing"
6.1
6.2
6.3
6.3.1
6.3.2
6.3.3
6.3.4
5.1
5.2
5.3
5.4
5.5
5.6
5.7
5.8
4.1
4.2
4.3
4.3.1
4.3.2
4.4
4.5
4.6
4.7
4.8
4.8.1
4.8.2
4.8.3
Conventions and Symbols used in these Operating Instructions
S Platform
L Platform
Operation Keys
Display Panel
Unpacking and Delivery Inspection
Installing the Components
Selecting the Location and Leveling the Balance
Selecting the Location
Leveling the Balance
Power Supply
Battery Operation
Transporting the Balance
Weighing Below the Balance
Adjustment (Calibration)
Fully Automatic Adjustment FACT
Manual Adjustment with Internal Weight
Manual Adjustment with External Weight
Switching the Balance On and Off
Performing a Simple Weighing
Zeroing / Taring
METTLER TOLEDO DeltaRange Balances
Switching Weight Units
Recall / Recall Weight Value
Weighing With the Weighing-in Aid
Print / Transmit Data
What is in the Menu
Menu Operation
Description of Menu Topics
Menu Mode
Basic Menu
Advanced Menu
Interface Menu
3
Table of contents
9
10
11
12
13
14
15
16
17
18
19
20
Application "Check Weighing"
Application "Statistics"
Application "Formulation" (Net Total Formulation)
Application "Totaling"
Application "Dynamic Weighing"
Application "Multiplication Factor Weighing"
Application "Routine Test"
Application "Diagnostics"
16.1
16.2
16.3
16.4
16.5
16.6
16.7
16.8
Repeatability Test
Display Test
Key Test
Motor Test
Balance History
Calibration History
Balance Information
Service Provider Information
Error and Status Messages
17.1
17.2
Error Messages
Status Messages
Cleaning and Service
18.1 Cleaning the Glass Draft Shield (0.1 mg and 1 mg Models)
Interface Specification
19.1
19.2
19.3
RS232C Interface
USB Device Interface
MT-SICS Interface Commands and Functions
Technical Data
20.1
20.2
20.2.1
20.2.2
20.2.3
20.2.4
20.2.5
20.2.6
20.3
20.3.1
20.3.2
20.3.3
20.3.4
20.3.5
General Data
Model-Specific Data
Balances with Readability of 0.1 mg, S Platform with Draft Shield
Balances with Readability of 1 mg, S Platform with Draft Shield
Balances with Readability of 0.01 g, S Platform
Balances with Readability of 0.1 g, S Platform
Balances with Readability of 0.1 g to 1 g, L Platform
Balances with Readability of 2 g to 5 g, L Platform
Dimensions
Balances with Readability of 0.1 mg, S Platform With Draft Shield
Balances with Readability of 1 mg, S Platform With Draft Shield
Balances with Readability of 0.01 g, S Platform
Balances with Readability of 0.1 g, S Platform
Balances with Readability of 0.1 g to 5 g, L Platform
4
21
22
23
Accessories and Spare Parts
21.1
21.2
Appendix
22.1
Index
Accessories
Spare Parts
Conversion Table for Weight Units
Table of contents
5
Introduction
1 Introduction
Thank you for choosing a METTLER TOLEDO balance.
The precision balances of the NewClassic line combine a large number of weighing possibilities with easy opera tion.
These operating instructions apply to all balance models MS-S and MS-L in the NewClassic line. However, the dif ferent models have different charcteristics regarding equipment and performance. Special notes in the text indicate where this makes a difference to operation.
1.1 Conventions and Symbols used in these Operating Instructions
Key designations are indicated by double angular parantheses (e.g. « »).
This symbol indicates press key briefly (less than 1.5 s).
This symbol indicates press and hold key down (longer than 1.5 s).
This symbol indicates a flashing display.
This symbol indicates an automatic sequence.
These symbols indicate safety notes and hazard warnings which, if ignored, can cause personal danger to the user, damage to the balance or other equipment, or malfunctioning of the balance.
This symbol indicates additional information and notes. These make working with your balance easier, as well as ensuring that you use it correctly and economically.
7
8
Safety Precautions
2 Safety Precautions
Always operate and use your balance only in accordance with the instructions contained in this manual. The instructions for setting up your new balance must be strictly observed.
If the balance is not used according to the manufacturer's Operating Instructions, protection of the balance may be impaired.
It is not permitted to use the balance in hazardous environments.
Use the MSxxKLIP balance model with Protection Class IP65 if: the balance is used in wet areas, wet claning is necessary or the balance is used in a dusty environment. Even with Pro tection Class IP65. Never flood the balance or immense it in liquid.
All other balance models may only be used in dry rooms.
Use only the Universal AC adapter delivered with your balance.
The L platform has a built-in power supply unit. Hazard of electric shock if the power cable is damaged! Check the power cable for damage regularly. Unplug the power cord immediately if the power cable is damaged.
Do not use sharply pointed objects to operate the keyboard of your balance! Although your bal ance is very ruggedly constructed, it is nevertheless a precision instrument. Treat it with corre sponding care.
Do not open the balance: It does not contain any parts which can be maintained, repaired, or replaced by the user. If you ever have problems with your balance, contact your METTLER TOLE
DO dealer.
Use only balance accessories and peripheral devices from METTLER TOLEDO; they are optimally adapted to your balance.
Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic
Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point speci fied for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.
Thank you for your contribution to environmental protection.
Overview
3 Overview
3.1 S Platform
3
3
5
7
0.1 mg
1
C
2 14 10
4
6
5
7
9
C
10 mg
1
2 14
10
6
7
5
1 mg
C
1 2 14 10 4
7
5
C
0.1 g
1
2
14 10
9
15
Name and Function of Components
1
Display
11 12 13 8
9
Safety foots (with 10 mg, 0.1 g S series models)
9
Overview
Name and Function of Components
2
Operation keys
3
Glass draftshield
4
Handle for operation of the draft-shield door
5
Weighing pan
6
Draft shield element
7
Level indicator
8
Kensington slot for anti-theft purposes
3.2 L Platform
10 Leveling foot
11 USB device interface
12 RS232C serial interface
13 Socket for AC Adapter
14 Model sticker (with approved models only)
15 Product label
3
4
5
1
10
2 5
5
Name and Function of Components
1
Display
2
Operation keys
3
Weighing pan
4
Level indicator
5
Leveling foot
6
USB device interface
6 7
9
11
8 5
7
RS232C serial interface
8
Power cord with country-specific plug
9
Security slot for anti-theft purposes
10 Model sticker (with approved models only)
11 Product label
10
Overview
3.3 Operation Keys
1
2
Weigh-Track
3
C
4
5 6 9 7 10 8
Key Functions
No.
Key
1
2
3
4
5
6
●
●
Press briefly (less than 1.5 s)
To change display resolution (1/10d func tion) while application is running
Press and hold (longer than 1.5 s)
no function
Enter and leave menu (Parameter settings) no function
●
●
●
●
●
●
●
●
Execute predefined adjusting (calibration) procedure
Printout display value
Printout active user menu settings
Transfer data
To navigate back (scroll up) within menu topics or menu selections (instead of using key no.7)
Decrease (numerical) parameters within menu and in applications
To navigate forth (scroll down) within menu topics or menu selections
Increase (numerical) parameters within menu and in applications no function no function
●
●
●
To select the weighing application
To select assigned F1 application and entering the parameter settings of applica tion.
Default F1 application assignment: Piece counting
11
Overview
No.
7
8
9
10
Key
●
●
●
●
●
●
●
●
●
Press briefly (less than 1.5 s)
With entries: scroll down
To navigate through menu topics or menu selections
To toggle between unit 1, recall value (if selected), unit 2 (if different from unit 1) and the application unit (if any)
To enter and leave menu selection (from / to menu topic)
To enter application parameter and switch to next parameter
To store parameter
Switch on
Zero/Tare
Cancel and to leave menu without saving
(one step back in the menu).
●
Press and hold (longer than 1.5 s)
To select assigned F2 application and entering the parameter settings of applica tion.
● Default F2 application assignment: Percent weighing
●
●
●
To select assigned F3 application and entering the parameter settings of applica tion.
Default F3 application assignment: Check weighing
Switch off no function
3.4 Display Panel
Application Icons Status Icons
12
Application Icons
Menu locked
Settings activated
Application "Weighing"
Application "Piece counting"
Application "Percent weighing"
Application "Check weighing"
Status Icons
Indicates stored value (Memory)
Indicates Net weight values
Adjustments (calibration) started
FACT activated
Weighing-in aid Status Icons
Application "Statistics"
Application “Formulation / Net-Total”
Application “Totaling”
Application “Dynamic weighing”
Application “Multiplication factor” n.a.
Service reminder
Acoustic feedback for pressed keys activated
Weighing range 1 (Dual Range models only)
Weighing range 2 (Dual Range models only)
Overview
Status Icons
Applications "Diagnostics" and “Routine Test”
Weight Value Field and Weighing-in aid
Indicates negative values
Unit Field
Indicates unstable values
Indicates calculated values
g
gram
kg
kilogram
mg
miligram
ct
carat
lb
pound
oz
ounze
Charge of battery: full, 2/3, 1/3, discharged
(Battery operated models only)
Brackets to indicate uncertified digits
(approved models only)
Marking of nominal or target weight
Marking of tolerance limit T+
Marking of tolerance limit T-
ozt
troy ounce
GN
grain
dwt
pennyweight
mom
momme
msg
mesghal
tlh
Hong Kong taels
tls
Singapore taels
tlt
Taiwan taels
tola
tola
bath
bath
13
Setting up the Balance
4 Setting up the Balance
The balance must be disconnected from the power supply when carrying out all setup and mounting work.
4.1 Unpacking and Delivery Inspection
a) Open the packaging and carefully remove all components.
b) Check the delivered items.
The standard scope of delivery contains the following items:
Components S platform
0.1 mg 1 mg
– Draft shield
Weighing pan with pan support
237 mm
165 mm
Ø 90 mm
127 x 127 mm
170 x 200 mm
190 x 226 mm
246 x 351 mm
Draft shield element
Pan support
Bottom plate
Protective cover
Universal AC adapter (country specific)
Mounted country specific power cable
Operating instruction (this document)
Quick Guide
EC declaration of conformity
–
–
–
–
–
–
–
–
–
–
–
–
0.01 g
–
–
–
–
–
–
–
L platform
0.1 g 0.1g/1g 2g/5g
– – –
– – –
–
–
–
–
–
–
–
–
–
–
–
–
– –
–
–
–
– – –
– –
–
14
Setting up the Balance
4.2 Installing the Components
1
C
6
5
4
3
2
Balances with readability of 0.1 mg, S platform with draft shield (237 mm)
Place the following components on the balance in the specified order:
Note: Push the side glass back as far as will go and grasp the draft shield (1) with both hands on the bars at the top. a) Turn draft shield lock (2) to position " "(unlock).
b) Place draft shield on the balance.
c) Turn draft shield lock to " " (lock) and place bot tom plate (3).
d) Place draft shield element (4) and weighing pan
(6) with pan support (5).
Note: Cleaning the draft shield see section "Mainte nance and cleaning".
1
C
5
4
3
2
Balances with readability of 1 mg, S platform with
draft shield (165 mm)
Place the following components on the balance in the specified order:
Note: Push the side glass back as far as will go and grasp the draft shield (1) with both hands on the bars at the top. a) Turn draft shield lock (2) to position " "(unlock).
b) Place draft shield on the balance.
c) Turn draft shield lock to " " (lock) and place bot tom plate (3).
d) Place weighing pan (5) with pan support (4).
Note: Cleaning the draft shield see section "Mainte nance and cleaning".
15
Setting up the Balance
C
C
3
2
1
●
●
Balances with readability of 10 mg, S platform
Place the following components on the balance in the specified order:
●
Pan support (1)
Weighing pan (2)
Draft shield element (3)
2
1
●
●
Balances with readability of 0.1 g, S platform
Place the following components on the balance in the specified order:
Pan support (1)
Weighing pan (2)
1
Balances with readability to 1 g, L platform
Place the weighing pan (1) on the balance.
16
1
Setting up the Balance
Balances with readability from 2 g, L platform
Place the weighing pan (1) on the balance.
4.3 Selecting the Location and Leveling the Balance
Your balance is a precision instrument and will thank you for an optimum location with high accuracy and dependability.
4.3.1 Selecting the Location
Select a stable, vibration-free position that is as horizontal as possible. The surface must be able to safely carry the weight of a fully loaded balance.
Observe ambient conditions (see Technical Data).
Avoid the following:
●
Direct sunlight
●
Powerful drafts (e.g. from fans or air conditioners)
● Excessive temperature fluctuations
4.3.2 Leveling the Balance
The balances have a level indicator and two (S Platform) or four (L Plat form) adjustable leveling feet to compensate for slight irregularities in the surface of the weighing bench. The balance is exactly horizontal when the air bubble is in the middle of the level glass.
Note: The balance should be leveled and adjusted each time it is moved to a new location.
Balances with S platform and readability of 0.1 mg and 1 mg
Adjust the two leveling feet appropriately until the air bubble comes to rest exactly in the middle of the glass:
Air bubble at "12 o'clock" turn both feet clockwise
17
Setting up the Balance
B
A
C
A
B
Air bubble at
Air bubble at
Air bubble at
"3 o'clock"
"6 o'clock"
"9 o'clock" turn left foot clockwise, right foot counterclockwise turn both feet counterclockwise turn left foot counterclockwise, right foot clockwise
Balances with S platform and readability of 10 mg and 0.1 g
a) Remove the clamps (A) for the safety feet by turning them outwards.
Note: Turn the clamps (A) outwards as far as they will go (~ 90°), so that the safety feet can move freely.
b) Now level the balance by turning both leveling screws (B) until the air bubble is in the inner circle of the level indicator (see procedure above).
c) Secure the safety feet by turning the clamps (A) inwards as far as they will go.
C
Balances with L platform
Align the balance horizontally by turning the leveling screws of the bal ance housing until the air bubble is in the inner circle of the level indica tor.
4.4 Power Supply
Your balance is supplied with an country-specific AC adapter or with a country-specific power cable. The power supply is suitable for all line voltages in the range: 100 - 240 VAC, 50/60 Hz (for exact specifications, see section
"technical data").
First, check the local line voltage is in the range 100 - 240 VAC, 50/60 Hz and whether the power plug fits your local power supply connection. If this is not the case, on no account con
nect the balance or the AC adapter to the power supply, but contact the responsible METTLER
TOLEDO dealer.
Important: Guide the cables so that they cannot become damaged or interfere with the weighing process! Take care that the AC adapter cannot come into contact with liquids!
Allow your balance to warm up for 30 minutes (0.1 mg models 60 minutes) to enable it to adapt itself to the ambient conditions.
18
Setting up the Balance
Connect the AC adapter to the connection socket on the back of your bal ance (see figure) and to the power line.
4.5 Battery Operation
Balances with a built-in rechargeable battery can, under normal operation conditions, work independently of the AC power line for about 8 hours. Immediately the AC power supply is interrupted e.g. by withdrawing the power cord plug or if there is a powerfailure, the balance switches automatically to battery operation. Once the AC power sup ply is restored, the balance reverts automatically to AC operation.
fully charged
2/3 charged
1/3 charged discharged
When the balance is operating on its batteries, the battery symbol in the display lights up. The number of segments that are lit is an indicator of battery condition (3 = fully charged, 0 = discharged). When the batteries are almost completely discharged, the battery symbol flashes.
Charging the built-in battery is indicated with filling up all 3 levels contin uosly. After the charging is finished the battery symbol is turned off. The battery is protected against overcharging, and the balance can therefore remain permanently connected with the AC power line.
Note: The built-in rechargeable battery can not be replaced by the user.
Please contact METTLER TOLEDO customer Service.
4.6 Transporting the Balance
Switch off the balance and remove the power cable and any interface cable from the balance. Refer to the notes in
Section "Selecting the location" regarding the choice of an optimal location.
Transporting Over Short Distances
For balances with a draft shield: Observe the following instructions to transport your balance over a short distance to a new location: Never lift the balance using the glass draft shield.
The draft shield is not sufficient fastened to the balance.
Transporting Over Long Distances
If you would like to transport or send your balance over long distances, use the complete original packaging.
4.7 Weighing Below the Balance
The balances are equipped with a hanger for carrying out weighings below the work surface (weighing below the balance).
19
Setting up the Balance
Attention:
● Do not place the balance on the pan support location bolt (0.1 mg and 1 mg models).
● Models with a glass draft shield: Carefully lift the draft shield from the weighing platform and put it aside.
Note:
● For below-the-balance weighing with the L Platform models, you will need hook 11132565 from the accessories range.
● Weighing below the balance is not possible with "MS-KL" models.
S Platform L Platform
A
B
A
B
a) Switch off the balance and remove the power cable and any interface cable from the balance.
b) Remove the draft shield element (10 mg models).
c) Remove the weighing pan and pan support.
d) Remove the bottom plate and unlock the draft shield (models with draft shield).
e) Remove the 2 screws (A) and the cover plate (B). The hanger is now accessible.
f) Then turn the balance to its normal position and simply reinstall all components in the reverse order.
4.8 Adjustment (Calibration)
To obtain accurate weighing results, the balance must be adjusted to match the gravitational acceleration at its location. Adjusting is necessary:
● before the balance is used for the first time.
● at regular intervals during weighing service.
● after a change of location.
4.8.1 Fully Automatic Adjustment FACT
Note: On models with FACT only.
The factory setting is fully automatic adjustment FACT (Fully Automatic Calibration Technologiy) with the internal weight (see also section "The Menu"). In this setting, you have no need worry about adjusting your balance.
The balance adjusts itself automatically:
● after the warm-up phase on connection to the power supply.
20
ADJ.INT
------
D: 0.00
%
ADJ DONE
0.00
g
Setting up the Balance
●
●
● when a change in the ambient conditions, e.g. the temperature could lead to a noticeable deviation in the mea surement.
on a predefined time. (see menu topic "FACT") time interval. (with approved models)
4.8.2 Manual Adjustment with Internal Weight
Note: On models with internal weight only (see technical data).
To obtain accurate results, the balance must be connected to the power supply for approxi mately 30 minutes (0.1 mg balances 60 minutes) in order to reach operating temperature before adjusting.
Requirement: To carry out this operation, in the menu topic "CAL"
(Adjustment) of advanced menu "ADJ.INT" must be selected.
a) Unload weighing pan b) Press « » to execute "Internal Adjustment".
The balance adjusts itself automatically. The adjusting is finished when the deviation and the message "ADJ DONE" appears briefly in the dis play. The balance returns to the last active application and is ready for operation.
Sample adjustment printout using internal weight:
¦ -- Internal adjustment - ¦
12:56 ¦ ¦ 21.Jan. 2009
¦
¦ METTLER TOLEDO
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
Balance Type
SNR:
Temperature
Diff
MS4002S
1234567890
22.5 °C
0.001 %
¦
¦
¦
¦
¦
¦
¦
Adjustment done
-----------------------
¦
¦
¦
4.8.3 Manual Adjustment with External Weight
Note: Because of certification legislation, the approved models cannot be adjusted with an external weight.
To obtain accurate results, the balance must be connected to the power supply for approxi mately 30 minutes (0.1 mg balances 60 minutes) in order to reach operating temperature before adjusting.
21
Setting up the Balance
ADJ.EXT
2000.00
g
-------
0.00
g
------
D: 0.001
%
ADJ DONE
0.00
g
Sample adjustment printout using external weight:
¦ -- External adjustment - ¦
¦ 21.Jan. 2009 12:56 ¦
¦
¦ METTLER TOLEDO
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦ Balance Type
SNR:
Temperature
Nominal
Actual
Diff
MS4002S
1234567890
22.5 °C
4200.00 g
4199.99 g
0.001 %
¦
¦
¦
¦
¦
¦
¦
¦
Requirement: To carry out this operation, in the menu topic "CAL"
(Adjustment) " of advanced menuADJ.EXT" must be selected.
a) Have required adjustment weight ready.
b) Unload weighing pan.
c) Press « » briefly to execute "External Adjustment". The required (pre defined) adjustment weight value flashes in the display.
d) Place adjustment weight in center of pan. The balance adjusts itself automatically.
e) When "0.00 g" flashes, remove adjustment weight.
The adjusting is finished when the deviation value and the message "ADJ
DONE" appears briefly in the display. The balance returns to the last active application and is ready for operation.
¦
¦
¦
¦
¦
¦
¦
Adjustment done
Signature
.......................
-----------------------
¦
¦
¦
¦
¦
¦
¦
22
Weighing Made Simple
5 Weighing Made Simple
This section shows you how to perform simple weighings and how you can accelerate the weighing process.
5.1 Switching the Balance On and Off
This section shows you how to perform simple weighings and how you can accelerate the weighing process.
On
Off
Switching On
a) Remove any load from weighing pan.
b) press the «On» key.
0.00
g
The balance performs a display test (all segments in the display light up briefly), "WELCOME", Software version, Maximum load and Readability appears briefly. (Startup "FULL" mode only)
The balance is ready for weighing or for operation with the last active application.
On
Off
SHUTOFF
Switching Off
Press and hold the «Off» key until "SHUTOFF" appears on the display.
Release the key.
After it has been switched off, your balance is in the standby mode. The display shows date, time, maximum load and readability. As your balance needs no warm-up time in the standby mode and is immediately ready for weighing.
When Quickstart "QUICK" (Advanced menu, topic "STARTUP") is selected: If you wish to perform a weighing, you need now only place the sample on the weighing pan and the balance immedi ately displays the result. There is no need to switch it on with the «On/Off» key.
Note:
● Quickstart is not possible with approved balances.
●
Standby mode is available by line powered balances only.
23
Weighing Made Simple
5.2 Performing a Simple Weighing
WEIGHING
0.00
g
1182.03
g
¢¢¢¢¢¢
1250.38
g
¢¢¢¢¢¢¢¢ a) If your balance is not in the weighing mode, press and hold the
« » key down until “WEIGHING“ appears in the display. Release the key. Your balance is in the weighing mode and set to zero.
b) Place weighing sample on the weighing pan.
c) Wait until the instability detector " " disappears and the stability beep sounds.
d) Read the result.
5.3 Zeroing / Taring
#
#
121.01
g
¢¢¢¢
0.01
0.00
g g
0.00
g
121.01
g
¢¢¢¢
Net
0.00
g
¢¢¢¢
Net
95.97
g
¢¢¢¢¢¢¢
Net
-
¢¢¢¢
01 g
Zeroing
a) Unload the balance.
b) Press « » to set the balance to zero. All weight values are measured in relation to this zero point (see menu topic "ZERO
RNG").
Note: Use the « » zeroing key before you start with a weighing.
Taring
If you are working with a weighing container, first set the balance to zero.
a) Place empty container on the balance. The weight is displayed.
b) Press « » to tare the balance.
"0.00 g" and "Net" appears in the display. "Net" indicates that all weight values displayed are net values.
Note:
● If the container is removed from the balance, the tare weight will be shown as a negative value.
● The tare weight remains stored until the « again or the balance is switched off.
» key is pressed
●
With METTLER TOLEDO DeltaRange balances (see following section), the fine range with its 10 times smaller display increments (depends on the model) is available again after every taring operation.
24
Weighing Made Simple
5.4 METTLER TOLEDO DeltaRange Balances
# 812.7
g
¢¢¢¢
0.00
g
¢¢¢¢
METTLER TOLEDO DeltaRange balances have a movable fine range with
10 times smaller display increments over their entire weighing range. In this fine range an additional decimal place always appears in the dis play.
The balance operates in the fine range
● after switching on.
● after every taring operation.
If the fine range is exceeded, the balance display automatically switches to coarser display increments.
5.5 Switching Weight Units
The « » key can be used at any time to toggle between weight unit
"UNIT 1","RECALL" value (if selected) and weight unit "UNIT 2" (if differ ent from weight unit 2).
22.00
g
¢¢¢¢
M
*
22.00
g
0.78
oz
¢¢¢¢
5.6 Recall / Recall Weight Value
Recall stores stable weights with an absolute display value bigger than 10d.
Requirement: The function "RECALL" must be activated in the menu.
a) Load weighing sample. The display shows weight value and stores stable value.
0.00
g
121.01
¢¢¢¢ g b) Remove weighing sample. When the weight is removed the Display shows zero.
0.00
g c) Press The « » key. The display shows last stored stable weight value for 5 seconds together with asterisk (*) and Memory (M) sym bols. After 5 seconds the display goes back to zero. This can be repeated unlimited times.
M
121.01
g
0.00
g
Delete last weight value
As soon a new stable weight value is displayed, the old recall value becomes replaced by the new weight value. When pressing the
« » key, the recall value is set to 0.
Note: If the power is switched off, the recall value is lost. The recall value can not be printed.
25
Weighing Made Simple
5.7 Weighing With the Weighing-in Aid
¢¢¢¢¢¢¢¢
0%
100%
The weighing-in aid is a dynamic graphic indicator which shows the used ammount of the total weighing range. You can thus recognize at a glance when the load on the balance approaches the maximum load.
5.8 Print / Transmit Data
Pressing the « » key transmits the weighing results over the interface e.g. to a printer or a PC.
26
The Menu
6 The Menu
6.1 What is in the Menu
The Menu allows you to match your balance to your specific weighing needs. In the menu you can change the settings of your balance and activate functions. The menu has 3 modes and contains 37 different topics, each of which allows you various selection possibilities.
Note: See Quick Guide for the graphical overview of the menu (Menu Map) with all setting possibilities.
Menu Mode "BASIC"
Topic Description
DATE
TIME
UNIT 1
UNIT 2
KEY BEEP
RESET
SET MENU
Setting the current date.
Setting the current time.
Specification of the 1st weight unit in which the balance should show the result.
Specification of the 2nd weight unit in which the balance should show the result.
Setting the key beep level.
Call up of the factory settings.
Selection of the menu mode "BASIC", "ADVANCED", "INT.FACE" or "OFF".
Menu Mode "ADVANCED"
Topic Description
ENVIRON.
CAL
FACT
FACT PRT.
DATE
DATE.FORM
TIME
TIME.FORM
RECALL
UNIT 1
UNIT 2
KEY BEEP
STAB.BEEP
STARTUP
SHUTOFF
BCKLIGHT
DISPLAY
AUTOZERO
SERV.ICON
ZERO RNG
LANGUAGE
ASSIGN:F1
ASSIGN:F2
ASSIGN:F3
DIAGNOSE
Matching the balance to the ambient conditions.
Settings for the type of adjustment (calibration).
Settings for fully automatic balance adjustment based on a selected time.
Switching the automatic FACT printout on or off.
Setting the current date.
Setting the date format.
Setting the current time.
Preselection of the time format.
Switching the application "Recall" for storing stable weights on or off.
Specification of the 1st weight unit in which the balance should show the result.
Specification of the 2nd weight unit in which the balance should show the result.
Setting the key beep level.
Setting the stability beep level.
Setting the mode which the balance powers up (full or standby).
Setting the time after which the balance should be switched off automatically.
Setting the time after which the display backlight should be switched off automatically.
Adjusting the brightness and contrast of the display.
Switching the automatic zero correction (Autozero) on or off.
Switching the service icon (service reminder) on or off.
Setting the zero limit of the zero/tare key.
Setting the preferred language.
Selection of assigned F1 key application and entering their parameter settings.
Selection of assigned F2 key application and entering their parameter settings.
Selection of assigned F3 key application and entering their parameter settings.
Starting a diagnostic application.
27
The Menu
Topic
RESET
SET MENU
Description
Call up of the factory settings.
Selection of the menu mode "BASIC", "ADVANCED", "INT.FACE" or "OFF".
Menu Mode "INT.FACE"
Topic Description
RS232
BAUDRATE
BIT/PAR.
HD.SHAKE
RS E.O.L.
RS CHAR
USB
USB E.O.L.
USB CHAR
INTERVAL
SET MENU
Matching the serial interface RS232C to a peripheral unit.
Setting the transfer speed of the serial interface RS232C.
Setting the character format (Bit/Parity) of the serial interface RS232C.
Setting the transferprotocol (Handshake) of the serial interface RS232C.
Setting the end of line format of the serial interface RS232C.
Setting the char set of the serial interface RS232C.
Matching the USB interface to a peripheral unit.
Setting the end of line format of the USB interface.
Setting the char set of the USB interface.
Selection of the time which the balance started an automatic printout.
Selection of the menu mode "BASIC", "ADVANCED", "INT.FACE" or "OFF".
6.2 Menu Operation
In this Section you will learn how to work with the menu.
¢¢¢¢
406.2
g
BASIC
Opening the Menu
Press the menu key « » briefly. The balance shows the current menu mode for 1 second and switches to the first menu topic.
DATE >>
DATE
SET MENU
BASIC
ADVANCED
SET MENU
Set Menu Mode
a) Press «–» key to select the topic "SET MENU". (Last Menu Topic. ) b) Press « ». The current menu mode appears.
c) Press « » repeatedly to change menu mode. (scrolling down/up
«+» / «–» keys) d) Press « » to confirm the selection.
DATE
TIME
Select Menu Topic
Press « » briefly. The next menu topic appears in the display. Each time the « » or the «+» key is pressed, the balance switches to the next menu topic; the «–» key to the previous menu topic.
28
The Menu
/>
ENVIRON.
STABLE
ENVIRON.
+ 01.
2
500
+ 01.
6
500
+ 01.6500
Change Settings in a Selected Menu Topic
The ">>" flashing symbol in the display indicates selectable options available.
a) Press « ».The display shows the current setting in the selected menu topic. Each time « » or «+» is pressed, the balance switch es to the next selection; press «–» to the previous selection. After the last selection, the first is shown again.
b) Press « », the balance executes the selected setting and returns to the menu topic.
Change Settings in a Submenu Selection
The same procedure as for menu topics.
Input Principle of Numerical Values
a) Press « » for input of numerical values.
b) Press « » to select a digit or a value (depends on the applica tion). The selected digit or the selected value is blinking.
c) For changing digits or values, press «+» to scrolling up or «–» to scrolling down. d) Press « » to confirm the input.
SAVE:YES
SAVE:NO
Saving Settings and Closing the Menu
a) Press « » briefly to leave menu topic.
b) Press « » to execute "SAVE:YES". Changes are saved.
c) Press « » to execute "SAVE:NO". Changes are not saved. To toggle between "SAVE:YES" and "SAVE:NO" press « ».
Cancel
For leaving menu topic or menu selection without saving press «C» (one step back in the menu).
Note: If no entry is made within 20 seconds, the balance reverts to last active application mode. Changes are not saved. If changes are made, the balance asks "SAVE:NO.
6.3 Description of Menu Topics
In this Section you will find information regarding the individual menu topics and the available selections.
29
The Menu
6.3.1 Menu Mode
"SET MENU" – Menu Mode
Selecting the menu mode.
"BASIC"
"ADVANCED"
"INT.FACE"
"OFF"
6.3.2 Basic Menu
The small "BASIC" menu for simple weighing is visible.
The full "ADVANCED" menu for all weighing settings is visible.
The menu topics for all interface parameter settings for peripheral devices e.g. printer are visible.
Menu Selection "OFF" is visible only.
"DATE" – Date
Setting the current date according to date format.
Note: A reset of the balance will not change this setting.
"TIME" – Time
Setting the current time according to time format
"SET TIME"
"+1H"
"-1H"
Enter the current time.
Set the current time forwards by 1 hour. (to adjust summer or win ter time).
Set the current time backwards by 1 hour. (to adjust summer or winter time)
Note: A reset of the balance will not change this setting.
"UNIT 1" – Weight Unit 1
Depending on requirements, the balance can operate with the following units (depend on the model)
● Only those weight units allowed by the appropriate national legislation are selectable.
●
●
With approved balances, this menu topic has a fixed setting and cannot be changed.
Conversion table for weight units see chapter Appendix.
Units:
g
1)
kg
2)
mg
3)
ct
Gram
Kilogram
Miligram
Carat
lb oz ozt
Pound
Ounce (avdp)
Ounce (troy)
GN
Grain
1)
factory setting
2)
not with 0.01 mg and 1 mg balances
3)
with 0.1 mg and 1 mg balances
4)
the Malaysian tael has the same value
dwt mom msg tlh tls
4)
tlt tola baht
Pennyweight
Momme
Mesghal
Tael Hong Kong
Tael Singapore
Tael Taiwan
Tola
Baht
30
The Menu
"UNIT 2" – Weight Unit 2
If it is required to show the weighing results in weighing mode in an additional unit the desired second weight unit can be selected in this menu topic (depends on the model). Units see "UNIT 1". Select "NO", if you do not want to use "UNIT 2".
Note: Only those weight units allowed by the appropriate national legislation are selectable.
"KEY BEEP" – Key Beep
This menu topic allows you to select the volume of the key beep. The according key beep is emitted during the set ting.
"MED"
"HIGH"
"OFF"
"LOW"
Medium level (Factory setting)
High level
Beep switched off
Low level
"RESET" – Reset Balance Settings
This menu topic allows you to cal-up of the factory settings.
To toggle between "YES?" and "NO?" press « ».
Note: A reset of the balance will not change the "DATE" and "TIME" settings.
6.3.3 Advanced Menu
"ENVIRON." – Environment Settings
This setting can be used to match your balance to the ambient conditions.
"STANDARD"
"UNSTABLE"
"STABLE"
Setting for an average working environment subject to moderate variations in the ambient conditions. (Factory setting)
Setting for an working environment where the conditions are contin uously changing.
Setting for an working environment which is practically free from drafts and vibrations.
"CAL" – Adjustment (calibration)
In this menu topic you can preselect the function of the « » key. Your balance can be adjusted with internal or external weights by pressing the « » key. If you have attached a printer to your balance, the data of the adjustment
(calibration) are printed out.
"ADJ.OFF"
"ADJ.INT"
"ADJ.EXT"
"200.00 g"
The adjustment is switched off. The « » key has no function.
Internal adjustment: adjustment is performed at a keystroke with the built-in weight (depend on the model, see technical data).
External adjustment: adjustment is performed at a keystroke with a selectable external weight.
Note: This function is not available for approved balances.
Defining the external adjustment weight: define the weight of the external adjustment weight (in grams).
Factory setting: depends on the model.
31
The Menu
"FACT" – Fully Automatic Adjustment
Fully automatic internal adjustment (calibration) FACT (Fully Automatic Calibration Technology) provides fully automatic balance adjustment based on temperature criteria and on preselected time. (depend on the model, see technical data)
"TIME"
"OFF"
"12:00"
Execute FACT (with selected time).
Specify the time for a fully automatic adjustment to take place every day.
Factory setting: 12:00 (according to time format)
The selected time for FACT is switched off.
"FACT PRT." – Protocol Trigger for Fact
This setting specifies whether an adjustment report should be printed automatically.
Note: This menu topic does not affect the printing of adjustments with an internal or external adjustment weight.
"OFF"
"ON"
Protocol switched off: if the balance adjusts automatically (FACT), a protocol is not printed out.
Protocol switched on: a record is printed out after every automatic adjustment of the balance (FACT).
Note: The protocol is printed out without a line for signatures.
"DATE" – Date
Setting the current date according to date format.
Note: A reset of the balance will not change this setting.
"DATE.FORM" – Date Format
This menu topic allows you to preselect the date format.
The following date formats are available:
"DD.MM.Y"
"MM/DD/Y"
"Y-MM-DD"
"D.MMM.Y"
"MMM D Y"
Display examples
01.01.2009
01/01/09
09-01-01
1.JAN.09
JAN 1 09
Factory setting: "DD.MM.Y"
Printing examples
01.01.2009
01/01/2009
2009-01-01
1.JAN.2009
JAN 1 2009
"TIME" – Time
Setting the current time according to time format
"SET TIME"
"+1H"
"-1H"
Enter the current time.
Set the current time forwards by 1 hour. (to adjust summer or win ter time).
Set the current time backwards by 1 hour. (to adjust summer or winter time)
Note: A reset of the balance will not change this setting.
32
The Menu
"TIME.FORM" – Time Format
This menu topic allows you to preselect the time format.
The following date formats are available:
"24:MM"
"12:MM"
"24.MM"
"12.MM"
Display examples
15:04
3:04 PM
15.04
3.04 PM
Factory setting: "24:MM"
"RECALL" – Recall
This menu topic allows you to switch the "RECALL" function on or off. When it is switched on recall stores the last stable weight if the absolute display value was bigger than 10d.
"ON"
"OFF"
"RECALL" switched on
"RECALL" switched off
Note: The recall value is displayed with an asterisk and cannot be printed.
"UNIT 1" – Weight Unit 1
Depending on requirements, the balance can operate with the following units (depend on the model)
●
Only those weight units allowed by the appropriate national legislation are selectable.
●
With approved balances, this menu topic has a fixed setting and cannot be changed.
● Conversion table for weight units see chapter Appendix.
Units:
g
1)
kg
2)
mg
3)
ct lb oz
Gram
Kilogram
Miligram
Carat
Pound
Ounce (avdp)
ozt
GN
Ounce (troy)
Grain
1)
factory setting
2)
not with 0.01 mg and 1 mg balances
3)
with 0.1 mg and 1 mg balances
4)
the Malaysian tael has the same value
dwt mom msg tlh tls
4)
tlt tola baht
Pennyweight
Momme
Mesghal
Tael Hong Kong
Tael Singapore
Tael Taiwan
Tola
Baht
"UNIT 2" – Weight Unit 2
If it is required to show the weighing results in weighing mode in an additional unit the desired second weight unit can be selected in this menu topic (depends on the model). Units see "UNIT 1". Select "NO", if you do not want to use "UNIT 2".
Note: Only those weight units allowed by the appropriate national legislation are selectable.
33
The Menu
"KEY BEEP" – Key Beep
This menu topic allows you to select the volume of the key beep. The according key beep is emitted during the set ting.
"MED"
"HIGH"
"OFF"
"LOW"
Medium level (Factory setting)
High level
Beep switched off
Low level
"STAB.BEEP" – Stability Beep
If the unstable symbol disapears, the stability beep becomes active. This menu topic allows you to preselect the volume of the stability beep.
"MED"
"HIGH"
"OFF"
"LOW"
Medium level (Factory setting)
High level
Beep switched off
Low level
"STARTUP" – Startup Mode
You can set your balance such that it either immediately starts from the standby mode when you load a weight or it must be switched on with the «ON/OFF» key after which it then performs a display test.
Note: This topic in not visible with approved balances.
"FULL"
"QUICK"
Start with display test: You must switch on the balance with the
«ON/OFF» key. After it has been switched on, it performs a display test for approx. 2 sec. in which all display elements lights up, it shows "WELCOME" ,software version, maximum load and read ability. The balance is ready for weighing. This is the Factory set
ting.
"Quickstart": The balance can be started directly from the standby mode and is immediately ready for weighing. You can load the weight in the standby mode and the balance immediately shows the current weighing result.
Note: Standby mode is available by line powered balances only.
"SHUTOFF" – Automatic Shutoff
If the automatic shutoff function is activated, the balance automatically switches itself off after a preselected time of inactivity (i.e. with no key being pressed or changes of weight occuring etc.) and is switched to the standby mode.
"A.OFF 10" min
"A.OFF –"
"A.OFF 2" min
"A.OFF 5" min
Automatic shutoff after 10 minutes of inactivity. (Factory setting)
Automatic shutoff not activated.
Automatic shutoff after 2 minutes of inactivity.
Automatic shutoff after 5 minutes of inactivity.
"BCKLIGHT" – Backlight
Under this menu topic, the display backlight can be switched off automatically. If the automatic switch-off is acti vated, the backlight will turn off automatically after the selected period of inactivity has elapsed. The backlight is reactivated when a key is pushed or the weight is changed.
"B.L. ON" Backlight is always on. (Factory setting)
34
The Menu
"B.L. 30" s
"B.L. 1" min
"B.L. 2" min
"B.L. 5" min
Automatic switch-off after 30 seconds inactivity.
Automatic switch-off after 1 minute inactivity.
Automatic switch-off after 2 minutes inactivity.
Automatic switch-off after 5 minutes inactivity.
"DISPLAY" – Display Settings
This menu topic allows you to adjust brightness and contrast of the display.
"BRIGHTN"
"50%"
"CONTRAST"
"75%"
To set the brightness in 1% steps.
Factory setting: 50%
To set the contrast in 1% steps.
Factory setting: 75%
"AUTOZERO" – Automatic Zero Setting
This menu topic allows you to switch the automatic zero setting on or off.
“ON“
"OFF"
"AUTOZERO" switched on (factory setting). The automatic zero set ting continuously corrects possible variations in the zero point that might be caused through small amounts of contamination on the weighing pan.
"AUTOZERO" switched off. The zero point is not automatically cor rected. This setting is advatageous for special applications (e.g.
evaporation measurements).
Note: With approved balances, this setting is not available.
"SERV.ICON" – Service Reminder
This menu topic allows you to switch the service icon " " on or off.
"ON"
"OFF"
Service icon " " switched on (factory setting).You will be informed after one Year or 8000 operating hours to call service for recalibra tion. This will be indicated by the flashing service icon.
Service icon " " switched off.
"ZERO RNG" – Zero Range
This menu topic allows you to set a zero limit for the « key will execute a zero. Above this limit the «
» key. Up and including this limit the «
» key will execute a tare.
"WEIGHT"
"4200g"
To set the upper limit of the zeroing range as weight in the defini tion unit of the balance.
»
"PERCENT"
"100%"
To set the upper limit of the zeroing range as a percent of the total range of the balance.
"LANGUAGE" – Language
Factory setting: Generally, the language of the destination country (if available) or English is set.
The following languages are available:
"ENGLISH"
"DEUTSCH"
English
German
"ITALIANO"
"POLSKI"
Italian
Polish
35
The Menu
"FRANCAIS"
"ESPANOL"
French
Spanish
"CESKY"
"MAGYAR"
Czech
Hungarian
"ASSIGN:F1" – Assign Application Key F1
At this menu topic you can assign an application to the "F1" key. The following applications are available
(depends on the model):
"F1:COUNT"
"F1:PERC.W"
"F1:CHECKW."
"F1:STAT."
"F1:FORM."
"F1:TOTAL."
"F1:DYN."
"F1:F.MUL"
Piece counting (Factory setting)
Percent weighing
Checkweighing
Statistics
Formulation / Net-Total
Totaling
Dynamic weighing
Multiplication factor
"ASSIGN:F2" – Assign Application Key F2
At this menu topic you can assign an application to the "F2" key. The following applications are available
(depends on the model):
"F2:PERC.W"
"F2:CHECKW."
"F2:STAT."
"F2:FORM."
"F2:TOTAL."
"F2:DYN."
"F2:F.MUL"
"F2:COUNT"
Percent weighing (Factory setting)
Checkweighing
Statistics
Formulation / Net-Total
Totaling
Dynamic weighing
Multiplication factor
Piece counting
"ASSIGN:F3" – Assign Application Key F3
At this menu topic you can assign an application to the "F3" key. The following applications are available
(depends on the model):
"F3:CHECKW."
"F3:STAT."
"F3:FORM."
"F3:TOTAL."
"F3:DYN."
"F3:F.MUL"
"F3:R.TEST"
"F3:COUNT"
"F3:PERC.W"
Checkweighing (Factory setting)
Statistics
Formulation / Net-Total
Totaling
Dynamic weighing
Multiplication factor
Routine test
Piece counting
Percent weighing
36
The Menu
"DIAGNOSE" – Diagnostics Application
At this menu topic you can start a diagnostic application. For more information see chapter application "Diagnos tics".
The following diagnostics are available:
"DISPLAY"
"KEYPAD T"
"CAL.MOT. T"
"BAL.HIST"
"CAL.HIST"
"BAL.INFO"
"PROVIDER"
"REPEAT.T"
Display test
Key test
Motor test (models with internal weights only)
Balance history
Calibration history
Balance information
Service provider information
Repeatability test (models with internal weights only)
"RESET" – Reset Balance Settings
This menu topic allows you to cal-up of the factory settings.
To toggle between "YES?" and "NO?" press « ».
Note: A reset of the balance will not change the "DATE" and "TIME" settings.
6.3.4 Interface Menu
"RS232" – RS232C Interface
At this menu topic you can select the peripheral device connected to the RS232C interface and specify how the data is transmitted.
"PRINTER"
"PRT.STAB"
"PC-DIR."
"HOST"
"PRT.AUTO"
"PRT.ALL"
"PRT.STAB"
"PRT.AUTO"
"PRT.ALL"
"SEND.OFF"
"SEND.STB"
Connection to a printer.
If the « » key is pressed, the next stable weight value will be printed. (Factory setting)
Every stable weight value will be printed, without pressing the « » key.
If the « » key is pressed, the weight value will be printed regard less of stability.
Connection to a PC: the balance can send data (as a Keyboard) to the PC used for PC applications e.g. Excel.
If the « » key is pressed, the next stable weight value will be sent followed by an enter. (Factory setting)
Every stable weight value will be sent followed by an enter, without pressing the « » key.
If the « » key is pressed, the weight value will be sent followed by an enter regardless of stability.
Connection to a PC, Barcode Reader etc.: the balance can send data to the PC and receive commands or data from the PC).
Send mode switched off.(Factory setting)
If the « » key is pressed, the next stable weight value will be sent.
37
The Menu
"SEND.CONT"
"SEND.AUTO"
"SEND.ALL"
"2.DISPLAY"
All weight value updates will be sent regardless of stability, without pressing the « » key.
Every stable weight value will be sent, without pressing the « » key.
If the « » key is pressed, the weight value will be sent regardless of stability.
Connection of an optional auxiliary display unit (transmission parameters cannot be selected).
"BAUDRATE" – Bauderate RS232C
This menu topic allows you to match the data transmission to different serial RS232C receivers. The baud rate
(data transfer rate) determines the speed of transmission via the serial interface. For problem-free data transmis sion the sending and receiving devices must be set at the same value.
The following settings are available:
600 bd, 1200 bd, 2400 bd, 4800 bd, 9600 bd, 19200 and 38400 bd. (default: 9600 bd)
Note:
● Not visible for 2nd display.
● Each device has separate settings.
"BIT/PAR." – Bit/Parity RS232C
At this menu topic you can set the character format for the attached RS232C serial peripheral device.
"8/NO"
"7/NO"
"7/EVEN"
"7/ODD"
8 data bits/no parity (Factory setting)
7 data bits/no parity
7 data bits/even parity
7 data bits/odd parity
Note:
● Not visible for 2nd display.
● Each device has separate settings.
"HD.SHAKE" – Handshake RS232C
This menu topic allows you to match the data transmission to different RS232C serial receivers.
"XON/XOFF"
"RTS/CTS"
"OFF"
Software handshake (XON/XOFF) (Factory setting)
Hardware handshake (RTS/CTS)
No handshake
Note:
●
Not visible for 2nd display.
● Each device has separate settings.
"RS E.O.L." – End of Line RS232C
At this menu topic you can set the "End of Line" character of the tansmitted data to different RS232C serial receivers.
"(CR)(LF)" <CR><LF> Carriage Return followed by Line feed (ASCII-Codes
013+010) (Factory setting)
38
The Menu
"(CR)"
"(LF)"
Note:
● Not visible for 2nd display.
● Each device has separate settings.
<CR> Carriage Return (ASCII-Code 013)
<LF> Line feed (ASCII-Code 010)
"RS CHAR" – Char Set RS232C
At this menu topic you can set the "Character Set" of the tansmitted data to different RS232C serial receivers.
"IBM/DOS"
"ANSI/WIN"
Char Set IBM/DOS (Factory setting)
Char Set ANSI/WINDOWS
Note:
● Not visible for 2nd display.
●
Each device has separate settings.
"USB" – USB Interface
At this menu topic you can select the mode of the "USB Device" interface and specify how the data is transmitted.
Note: this port is not usable for printers or displays.
"USB"
"SEND.OFF"
"SEND.STB"
"SEND.CONT"
"SEND.AUTO"
"SEND.ALL"
Select the mode of the "USB Device" interface
Send mode switched off (Factory setting)
If the « » key is pressed, the next stable weight value will be sent.
All weight value updates will be sent regardless of stability, without pressing the « » key.
Every stable weight value will be sent, without pressing the « » key.
If the « » key is pressed, the weight value will be sent regardless of stability.
"USB E.O.L." – End of Line USB
At this menu topic you can set the "End of Line" character of the tansmitted data to USB device.
"(CR)(LF)"
"(CR)"
"(LF)"
<CR><LF> Carriage Return followed by Line feed (ASCII-Codes
013+010) (Factory setting)
<CR> Carriage Return (ASCII-Code 013)
<LF> Line feed (ASCII-Code 010)
"USB CHAR" – Char Set USB
At this menu topic you can set the "Character Set" of the tansmitted data to USB device.
"ANSI/WIN"
"IBM/DOS"
Char Set ANSI/WINDOWS (Factory setting)
Char Set IBM/DOS
39
The Menu
"INTERVAL" – Print Key Simulation
At this menu topic you can activate a simulation of the « » key. "INTERVAL" simulates a print key press every x seconds.
Range:
0 sec:
0 to 65535 seconds disables the print key simulation
Factory setting: 0 sec
40
Application "Piece Counting"
7 Application "Piece Counting"
The "Piece Counting" application allows you to determine the number of pieces put on the weighing pan.
Requirement: The function "COUNTING" must be assigned to an «Fx» key (see advanced menu topic "ASSIGN:Fx", factory setting: F1).
COUNTNG
Activate function "COUNTING" by pressing and hold the appropriate assigned «Fx» key (factory setting: F1).
A
A B
C
FIX 10
PCS
---------
FIX 10
PCS
10
PCS
Piece counting needs first a reference weight, there are 3 possibilli ties:
A
Setting the reference by number.
B
Setting the reference for 1 piece in weighing mode.
C
Setting the reference for 1 piece in manual mode.
Setting possibility
A
Setting the reference by number
a) Select a number of reference pieces by scrolling with « ». Possible numbers are 5, 10, 20 and 50. b) Press « » to tare. If using: place empty container on the weighing pan first or tare again.
c) Add the selected number of reference pieces to container.
d) Press « » to confirm.
B
1
PCS.WGT
0.00
g
Setting possibility
B
Setting the reference for one piece in weighing mode
a) Select "PCS.WGT" by scrolling with « ».
b) Press « » to tare. If using: place empty container on the weighing pan first or tare again.
c) Add one reference piece to container. The weight of one piece is dis played.
d) Press « » to confirm.
2.74
g
---------
1
PCS
41
Application "Piece Counting"
C
PCS.WGT
*
2.68
g
*
2.74
g
Setting possibility
C
Setting the reference for one piece in manual mode
a) Select "PCS.WGT" by scrolling with « ».
b) Press « » to confirm. c) Enter the final reference one piece weight by scrolling up («+» key) or down («-» key). d) Press « » to confirm.
0
PCS
,
2.
2.
74
74 g g
Switching between manual mode and weighing mode
Press « » to switch between manual and weighing mode.
Note: By switching from weighing mode to manual mode the weight val ue will be transferred and can be changed manually.
Note: If you do not press a key for 20 seconds, the balance returns to the previous active application. Press «C» to cancel.
On completion of the setting procedure, your balance is ready for piece counting.
75
812.7
PCS g
Switching between piece count and weight display.
You can use the « » key at any time to switch the display between piece display, weighing unit "UNIT 1", "RECALL" value (if activated) and weighing unit "UNIT 2" (if different from "UNIT 1").
Note:
● The "RECALL" value is displayed with an asterisk (*) and icon "M" and can not be printed.
● Take into account minimum values: min. reference weight = 10d (10 digits), min. piece weight = 1d (1 digit)!
●
The current reference weight remains stored until the reference setting is changed.
42
Application "Percent Weighing"
8 Application "Percent Weighing"
The "Percent Weighing" application allows you to check a sample weight as percentage to a reference target weight.
Requirement: The function "PERCENT" must be assigned to an «Fx» key (see advanced menu topic "ASSIGN:Fx", factory setting: F2).
PERCENT
Activate function percent weighing "PERCENT" by pressing and hold the appropriate assigned «Fx» key (factory setting: F2).
A B
Percent weighing needs first a reference weight that should corre spond to 100%, there are 2 possibillities:
A
Setting the reference in manual mode (enter 100%).
B
Setting the reference in weighing mode (weigh 100%).
A
SET 100
%
*
175.28
g
Setting possibility
A
Setting the reference by manual mode(enter 100%) a) Press « » to activate manual mode.
b) Select the reference target weight (100%) by scrolling up («+» key) or down («–» key). Progressing speed by press and hold.
c) Press « » to confirm.
0.0
%
B
SET 100
%
0.00
g
Setting possibility
B
Setting the reference by weighing mode(weigh 100%) a) Press « » to tare the balance and to activate the weighing mode. If using: place empty container on the weighing pan first or tare again.
b) Load the reference weight (100%).
c) Press « » to confirm.
175.28
g
I00.0
%
¢¢¢¢
,
175.28
175.28
g g
Switching between manual mode and weighing mode
Press « » to switch between manual and weighing mode.
Note: By switching from weighing mode to manual mode the weight val ue will be transferred and can be changed manually.
Note:
If you do not press a key for 20 seconds, the balance returns to the previous active application.
On completion of the weighing-in procedure, your balance is ready for percent weighing.
43
Application "Percent Weighing"
,
73.17
%
128.25
g
128.25
g
128250 mg
Switching between percent and weight display
You can use the « » key at any time to switch the display between percent display, weighing unit "UNIT 1", "RECALL" value (if activated) and weighing unit "UNIT 2" (if different from UNIT 1).
Note:
●
The recall value is displayed with an asterisk (*) and can not be printed.
●
The current set weight remains stored until it is redetermined.
44
Application "Check Weighing"
9 Application "Check Weighing"
The "Check weighing" application allows you to check the deviation of a sample weight within a tolerance limit to a reference target weight.
Requirement: The function "CHECK" must be assigned to an «Fx» key (see advanced menu top ic "ASSIGN:Fx", factory setting: F3).
CHECK
Activate function "CHECK" by pressing and hold the appropriate assigned
«Fx» key (factory setting: F3).
1A
1A
1B
2
SET NOM .
10.00
g
20.28
g
Step 1: Check weighing needs first a reference weight that should correspond to the nominal weight, there are 2 possibillities:
1A
Setting the reference in manual mode. (enter nominal weight)
1B
Setting the reference in weighing mode. (weigh nominal weight)
Step 2: Check weighing needs the upper and lower limits:
2
Setting the upper and lower limits.
Setting possibility:
1A
Setting the reference by manual mode (enter nominal weight) a) Press « » to activate manual mode.
b) Select the reference target weight by scrolling up («+» key) or down
(«–» key). Progressing speed by press and hold.
c) Press « » to confirm the nominal weight.
1B
SET NOM .
0
.
00 g
20
.
28 g
Setting possibility:
1B
Setting the reference by weighing mode (weigh nominal weight) a) Press « » to tare the balance and to activate the weighing mode. If using: place empty container on the weighing pan first or tare again.
b) Load the nominal weight.
c) Press « » to confirm the nominal weight.
10.00
g
10.00
g
Switching between manual mode and weighing mode
Press « » to switch between manual mode and weighing mode.
Note: By switching from weighing mode to manual mode the weight val ue will be transferred and can be changed manually.
45
Application "Check Weighing"
2
SET T+/-
%
2.50
%
20.28
g
Step 2:
2
Setting the upper and lower limits (in percentage): a) Press « » to start setting.
b) Press « » to confirm the default limit of +/- 2.5 % or enter the limit value by scrolling up («+» key) or down («–» key). Press « » to confirm the limits.
Note:
● If you do not press a key for 20 seconds, the balance returns to the previous active application. Press «C» to cancel.
● The nominal weight must be at least 10 digit.
On completion of the setting procedure, your balance is ready for checkweighing.
20.15
g
Weighing-in-Aid
The Weighing-in-Aid helps you quickly determine the position of the sam ple weight regarding the tolerance.
1 Lower limit
2 Target weight
3 Upper limit
46
Application "Statistics"
10 Application "Statistics"
The "Statistics" application allows you to generate statistics of weighing values. 1 to 999 val ues are possible. Requirement: The function "STAT" must be assigned to an «Fx» key (see advanced menu topic "ASSIGN:Fx"). Connect a printer or a PC if present.
STAT
CLEAR M?
No
M .CLEARED
0.00
g
46.36
g
- 1 -
0.00
g
999
Displayed results:
a) Press « » to show the next statistical value.
b) Press «C» to cancel displaying results and to continue weigh ing next sample.
a) Activate function "STAT" by pressing and hold the appropriate assigned «Fx» key.
b) To continue the last statistics press « ». For a new statistical eval uation press « » to select "Yes" and press « » to clear the memory.
Note: If the memory is already cleared (sample counter is 0) the memory clear question will be not displayed.
Weighing the first sample weight:
a) Press « » to zero/tare the balance if needed.
b) Load the first sample weight.
c) Press « ». The display shows the samlple count "- 1 -" and the current weight is stored as sample and the weight is printed out.
Note: When the sample counter is displayed you may press «C» to undo this sample.
d) Unload the first sample weight.
Weighing further sample weights:
The same procedure as for the first sample weight.
●
1...999 samples are possible.
●
The next value will be accepted if the sample weight is in the range
70% –130% of the current average value and the dynamic weight has changed at least 100 digits. "OUT OF RANGE" will be displayed if the sample is not accepted.
Results:
Press « », the results are displayed and printed.
number of samples
N average
X standard deviation
S.DEV
relative standard deviation
S.REL
lowest value (minimum)
MIN
0.5 seconds
5
50. 712 g
3. 364 g
6.63
%
53.45
g
47
Application "Statistics"
Printout:
3
4
1
2
5 n x s dev s rel
Min.
Max.
Diff
Sum
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
-- Statistics --
21.Jan. 2009
Balance Type
SNR
12:56
MS4002S
1234567890
¦
¦
¦
¦
¦ ----------------------¦
46.36 g
55.81 g
47.49 g
53.28 g
49.71 g
5
50.712 g
3.364 g
6.63 %
46.36 g
55.81 g
9.45 g
235.56 g
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦ ----------------------¦
highest value (maximum)
MAX different between the minimum and the maximum sum of all values
DIFF
SUM
53.45
g
53.45
g
53.45
g
48
Application "Formulation" (Net Total Formulation)
11 Application "Formulation" (Net Total Formulation)
The "Formulation" (Net Total) application allows you to
● weighing in (add and store) up to 999 individual component weights and displays the total.
If a printer is connected, the component weights are printed individually and as a total.
● tare and store up to 999 container weights and displays the total. If a printer is connected, the tare weights are printed out individually and as a total.
Requirement: The function "FORMULA" must be assigned to an «Fx» key (see advanced menu topic "ASSIGN:Fx"). Connect a printer or a PC if present.
FORMULA
CLEAR M?
No
M .CLEARED
a) Activate function formulation "FORMULA" by pressing and hold the appropriate assigned «Fx» key.
b) Press « » to continue formulation weighing. For a new formulation press « » to select "Yes" and press « » to clear the memory.
Note: If the memory is already cleared (sample counter is 0) the memory clear question will be not displayed.
#
#
0.00
g
10.33
g
-T1-
0.00
Net g
Tare container (if used): a) Press « » to zero the balance if needed.
b) Place the empty container on the weighing pan.
c) Press « » . The container is tared and the tare count "- T1 -" is displayed and the tare weight is printed.
Note: If you pre-tare via MT-SICS (e.g. bar code reader) "- PT1 -" is dis played.
Net
Net
Weighing the first component weight:
a) Load the first component weight.
b) Press « ». The display briefly shows the component count "- 1 -" , the current weight is stored as sample and the component weight is printed. The display is set back to zero.
Weighing further component weights:
The same procedure as for the first component weight with the same or new container).
●
1...999 sample values are possible.
● max 999 tare values are possible.
Results:
Press « », the results are displayed and printed.
49
Application "Formulation" (Net Total Formulation)
Displayed results:
a) Press « » to show the next statistical value.
b) Press «C» to cancel displaying results and to continue weigh ing next component.
Printout:
¦
¦
¦
¦
-- Formulation --
21.Jan.2009
Balance Type
SNR:
12:56
MS4002S
1234567890
¦
¦
¦
¦
¦ ----------------------¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
1 T
1 N
2 N
2 T
3 N
.
.
n
T Total
G Total
10.33 g
8.85 g
9.23 g
10.84 g
7.43 g
999
452.76 g
546.79 g
N Total 94.03 g
¦ ----------------------- ¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
number of samples
N sum of all tare values (T and
PT) sum of all component gross weight values sum of all component net weight values
T.TOTAL
G.TOTAL
N.TOTAL
0.5 seconds
8
452.76
546.79
94.03
g g g
50
Application "Totaling"
12 Application "Totaling"
The "TOTALING" application allows you to weigh in different samples to add their weight values and to totalize them. 1 to 999 samples are possible.
Requirement: The function "TOTALING" must be assigned to an «Fx» key (see advanced menu topic "ASSIGN:Fx").
TOTALING
CLEAR M?
No
M .CLEARED
a) Activate function "TOTALING" by pressing and hold the appropriate assigned «Fx» key.
b) For a new totaling evaluation press « » briefly to enter "Yes" and press « » briefly to clear the memory.
Note: If the memory is already cleared (sample counter is 0) the memory clear question will be not displayed.
c) Press « » to zero the balance.
0.00
g
¢¢
0.00
g
46.36
g
¢¢¢¢¢¢
- 1 -
25.46
g
¢¢¢¢¢¢
- 2 -
999
Displayed results:
a) Press « » briefly to show the totalized value.
b) Press «C» briefly to cancel.
Weighing in the sample weight:
a) If using a container: place empty container on the weighing pan and press « » to zero the balance.
Note: The « the menu setting.
» key will always execute a zero, regardless of b) Load the first sample weight.
c) Press « ». The display shows the sample count "- 1 -" and the current weight is stored. The balance is set to zero.
Note: When the sample counter is displayed you may press «C» to undo this sample.
d) Unload the second sample weight. e) Press « ». The display shows the sample count "- 2 -" and the current weight is stored. The balance is set to zero.
Weighing in further sample weights:
● 1...999 samples are possible.
Results:
Press « », the results are displayed and printed.
number of samples
N totalized value
TOTAL
0.5 seconds
879
8789.79
g
Printout:
¦ ---- Totaling ---¦
51
Application "Totaling"
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
21.Jan. 2009
Balance Type
SNR:
12:56
MS1602S
1234567890
¦
¦
¦
¦ ----------------------¦
.
.
5
6
3
4
1
2
.
n
Total
46.36 g
55.81 g
47.49 g
53.28 g
49.71 g
53.93 g
879
8789.79 g
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦ ----------------------¦
52
Application "Dynamic Weighing"
13 Application "Dynamic Weighing"
The "Dynamic Weighing" application allows you to determine the weights of unstable samples or to determine weights under unstable ambient conditions. The balance calculates the weight as the average of a number of weighing operations over a defined time.
Requirement: The function "DYNAMIC" must be assigned to an «Fx» key (see advanced menu topic "ASSIGN:Fx").
Note: "Switching Units" and "RECALL" Functions are not available in this Application.
DYNAMIC
Activate function "DYNAMIC" by pressing and hold the appropriate assigned «Fx» key.
MOD. AUTO
MOD. MAN
TIME 3 s
TIME 5 s
1
Setting "Auto Start" or "Manual Start":
a) Press « » to select the mode:
– "Auto Start ""MOD AUTO" (default value). The weighing starts automatically on relative stability. However, the weighing sample must weigh at least 5 grams. For weighing samples below 5 g the weighing must be started manually.
– "Manual Start" "MOD MAN" b) Press « » to confirm the selection.
2
Setting the weighing time:
a) Press « » to select one of the available time intervals: 3 (default value), 5, 10, 20, 60 and 120 seconds.
b) Press « » to confirm the selected time interval.
Note: If you do not press a key for 20 seconds, the balance returns to the previous active application.
Your balance is now ready for dynamic weighing:
53
Application "Dynamic Weighing"
MAN.STRT
13.39
g
TIMER 5
TIMER 1
13.45
g a) Press « » to zero if needed.
b) If using a container: place empty container on weighing pan and press « » to tare the balance.
c) Load sample weight.
d) – If you have selected function "Manual Start" "MAN.STRT", press
« » to start the weighing.
– If you have selected function "Auto Start" "AUTO.STRT", the weigh ing starts automaticly on relative stability. For weighing samples below 5 g the weighing must be started manually by pressing « ».
e) Read off result. The result of the dynamic weighing is displayed with an asterisk (* = calculated value).
f) Unload sample weight.
g) "Manual Start" only, press « weighing.
» to zero and confirm the
Note:
● The remaining weighing time (in seconds) is displayed continously.
You can cancel the countdown by pressing «C».
●
The weight value remains in the display until the sample weight is removed from weighing pan and tare key « » is pressed
("Manual Start" only).
54
Application "Multiplication Factor Weighing"
14 Application "Multiplication Factor Weighing"
The "Multiplication Factor Weighing" application allows you to multiply the weight value (in grams) by a predefined factor (result = factor * weight) and have it calculated to a predefined number of decimal places.
Requirement: The function "FACTOR M" must be assigned to an «Fx» key (see advanced menu topic "ASSIGN:Fx").
FACTOR M
Activate function "FACTOR M" by pressing and hold the appropriate assigned «Fx» key.
SET F.MUL
0001.0000
Fac
0009.8
100
Fac
1
Setting the factor value:
Either the factor 1 appears as default value or the factor that was saved most recently.
a) Press « » to execute "SET F.MUL".
b) Press « » to select a digit. The selected digit is blinking.
c) For changing digits, press «+» to scrolling up or «–» to scrolling down. d) Press « » to confirm the selected factor (no automatic accep tance).
SET STEP
001.0
Stp
005.0
Stp
2
Setting the step value:
"SET STEP" appears in the display, and the program changes automati cally to allow the display increments to be entered. The smallest possible display increment appears as default value, or the last value that was saved.
a) Press « » to execute "SET STEP".
b) Press « » to select a digit. The selected digit is blinking.
c) For changing digits, press «+» to scrolling up or «–» to scrolling down. d) Press « » to confirm the selected step (no automatic acceptance).
Note: The allowed range for the step depends on the multiplication factor and the resolution of the balance. If it is outside the allowed range the error message "STEP OUT OF RANGE" will be displayed.
Note: If you do not press a key for 20 seconds, the balance returns to the previous active application. Press «C» to cancel.
On completion of the setting procedure, your balance is ready for multiplication factor weighing.
55
Application "Multiplication Factor Weighing"
0
9810
1000.02
g
Weighing procedure
a) Press « » to zero/tare.
b) Load sample weight on weighing pan.
c) Read the result. The appropriate calculation is then made using the weight of sample and the selected factor, the result being displayed with the selected display step.
Note: No units are displayed. d) Unload sample weight.
Toggeling between displaying the calculated value and the measured weight:
You can use the « » key to toggle between the calculated Value, weight value "UNIT 1", "RECALL" value (if selected) and weight value
"UNIT 2" (if different from "UNIT 1").
56
Application "Routine Test"
15 Application "Routine Test"
The "Routine Test" application allows you to determine the sensitivity of the balance. More about periodic sensitivity tests (routine tests) see: GWP
®
(Good Weighing Practice) on
www.mt.com/gwp.
GWP gives clear recommendation for routine testing:
● how should I test my balance?
●
● how often?
where can I reduce efforts?
More about test weights see www.mt.com/weights.
Requirement:
●
The function "R.TEST" must be assigned to «F3» key (see advanced menu topic
"ASSIGN:F3").
● It is recommended to connect a printer or a PC to the balance for showing the results.
R.TEST
USE TARE
5000.00
g a) Activate function "R.TEST" by pressing and hold the assigned «F3» key.
b) Select "No" (no tare weight used).
If a tare weight is used during the test select "Yes" (use a tare weight). To toggle between "Yes" and "No" use « . c) Press « » to confirm the selection.
Note:
● It is recommended to test the sensitivity without tare load. (factory setting "No").
● If using tare: Make shure that tare weight plus test weight is not exceeding max. load.
Setting the reference test weight value
The default value of the testweight: Next smaller OIML weight than the maximum load of your balance according to the GWP
®
recommenda tion.
a) For changing the value, press «+» to scroll up or «–» to scroll down.
Progressing speed by press and hold.
b) Press « » to confirm the value.
2.50
g
Setting the Control Limit
The default value of the control limit:
Test weight x weighing process tolerance / 2
Example: 5000 g x 0.1% / 2 = 2.50 g.
a) For changing the value, press «+» to scroll up or «–» to scroll down.
Progressing speed by press and hold.
b) Press « » to confirm the value.
57
Application "Routine Test"
1.25
g
Setting the Warning Limit
The default value of the warning limit:
Warning limit = control limit / safety factor
Example: 2.5 g / 2 = 1.25 g.
a) For changing the value, press «+» to scroll up or «–» to scroll down.
Progressing speed by press and hold.
b) Press « » to confirm the value.
Note: The default values of control limit and the warning limit are evaluated according the GWP recommendation.
These are based under the assumption that the weighing process tolerance is 0.1% and the safety factor is 2.
PRESS EN
5000.00
g
On completion of the setting procedure, your balance is ready for the routine test procedure.
Note: The test weight must be acclimatized to the ambient temperature of the balance.
a) Press « » to start the test.
PRESS EN b) Follow the instructions on the display. If the test weight value is flashing: Load the test weight (displayed value).
The printout starts after the weighing pan is unloaded.
Exit the current test procedure:
Press and hold « », «F1», «F2» for executing a new application or
«F3» to restart "Routine Test".
Printout:
¦ -- Routine Test --
¦ 21.Jan. 2009
¦
12:56 ¦
¦
¦ METTLER TOLEDO
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
Balance Type
SNR:
Sensitivity:
Test weight
Value
Warning L.
Control L.
Warning L.
Control L.
Signature
MS6002S/01
1234567890
5000.00 g
5000.11 g
1.25 g
2.50 g
OK
OK
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
........................
------------------------
¦
¦
¦
¦
58
Application "Routine Test"
What if Warning Limit or Control Limit are "FAILED"?
The "SOP for Periodic Sensitivity Tests (Routine Tests)" is providing information about measures when routine tests failed. Find a download version of these SOPs on www.mt.com/gwp, link "GWP
®
The Program / Routine Opera
tion".
Content of SOP:
●
Preparation
●
●
Test procedure
Evaluation
●
•
•
Deviation
If Warning Limit "FAILED"
If Control Limit "FAILED"
59
Application "Diagnostics"
16 Application "Diagnostics"
The "Diagnostics" application allows you to to carry out predefined diagnostics tests and to view or print predefined sets of balance information. This diagnostics tool helps you finding errors faster and more efficient.
Requirement: A printer or a PC is connected to the balance for showing the results.
a) Activate "ADVANCED" menu. (see section menu operation) b) Activate function "DIAGNOSE" by pressing « ».
c) Use « » to select appropriate tests.
16.1 Repeatability Test
The repeatability test allows you to repeat tests with internal weight for a given number of times.
Note: On models with internal weights only.
a) Press « » to activate repeatability test "REPEAT.T". "R.TST. 10" appears on the Display.
b) Enter the number of times (blinking) the test must be repeated by pressing «+» or «–». Possible values are 5,
10 (default), 20, 50, 100 times.
c) Press « » to start the test. The message "RUNNING REPEAT TEST" is displayed till the tests are completed.
d) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOSE".
Sample Printout:
¦ -- Repeatability Test -¦
¦ 21.Jan. 2009 11:34 ¦
¦
¦ METTLER TOLEDO
¦
¦
¦
¦
¦
¦
Balance Type
SNR:
SW:
Temperature
No. of tests
MS6002S/01
1234567890
V1.00
21.3 °C
10
¦
¦
¦
¦
¦
¦
¦ -----------------------¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
1.Value
1. Time
1. Temp.
2. Value
2. Time
2. Temp.
·
·
·
2813.00 g
00:00:00
21.3 °C
2813.01 g
00:00:04
21.3 °C
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦ -----------------------¦
60
Application "Diagnostics"
¦
¦
¦
¦
¦
¦
¦
¦
Max.
Min.
x
s Dev.
Max Temp.
Min Temp.
Mean Temp.
Total Time
2813.01 g
2813.00 g
2813.005 g
0.004 g
21.3 °C
21.3 °C
21.3 °C
00:00:44
¦
¦
¦
¦
¦
¦
¦
¦
¦ -----------------------¦
Examples:
Repeatability test is a tool to do functional check with the balance. It may be performed:
●
•
To check function of balance
during installation to store print out with installation documents.
•
• after preventative maintenance to store print out with installation maintenance report.
when remarkable decrease of weighing performance occurs to email/fax print out to service support provider for diagnose purposes.
● To develop the optimal environment settings (see menu topic "ENVIRON.").
Perform repeatability test each with "STABLE", "STANDARD" and "UNSTABLE" setting. The setting with the fastest total time suits best for the existing environmental conditions.
16.2 Display Test
The display test allows you to test the display of the balance.
a) Press « » to start "DISPLAY".
All possible segments and icons on the display will illuminate.
b) Press « » to print the test information.
c) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOSE".
¦
¦
¦
¦
¦
Sample Printout:
¦ -- Display Test -¦
11:34 ¦ ¦ 21.Jan. 2009
¦
¦ METTLER TOLEDO
¦
¦
Balance Type
SNR:
SW:
Display Test
MS204S
1234567890
V1.00
DONE
¦
¦
¦
¦
¦
¦ ----------------------¦
16.3 Key Test
The key test allows you to test the keys of the balance.
61
Application "Diagnostics" a) Press « » to start "KEYPAD T".
b) The message "KEY TEST - PRESS KEY TO BE TESTED" is displayed scrolling during the duration of the key test.
Press every Key briefly. Each press of a key beeps and echoes with "OK” on the display.
c) Second press «C» key to print the test information. The test procedure will be cancelled and the balance will return to the topic "DIAGNOSE". If a key has not been tested before printing, then the test results will be indicat ed with a "----" line.
Sample Information Displayed:
Key
« »
« »
« »
« »
«–»
«+»
« »
« »
«C»
« »
Sample Printout:
¦ -- Key Test --
¦ 21.Jan. 2009
¦
¦ METTLER TOLEDO
¦
11:34 ¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦ Balance Type
SNR:
SW:
1/10 Key
Menu Key
Cal Key
Print Key
Plus Key
Toggle Key
Enter Key
Cancel Key
0/T Key
¦ ------------------------
MS204S
1234567890
V1.00
OK
OK
OK
OK
OK
OK
OK
OK
OK
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
Display
1/10 D OK
MENU OK
CAL OK
PRINT OK
MINUS OK
PLUS OK
TOGGLE OK
ENTER OK
C OK
0/T OK
16.4 Motor Test
The motor test allows you to test the calibration motor of the balance.
Note: On models with internal weight only.
a) Press « » to start "CAL.MOT. T".
"RUNNING" is displayed during the Motor Test. A motor test is deemed successful when all the motor positions have been successfully tested. At the end of the test, the test information will be printed.
62
Application "Diagnostics" b) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOSE".
c) Press « » for printout.
Sample Printout:
¦ -- Motor Test -¦
11:34 ¦ ¦ 21.Jan. 2009
¦
¦ METTLER TOLEDO
¦
¦
¦
¦
¦
¦
¦
Balance Type
SNR:
SW:
Motor Test
MS204S
1234567890
V1.00
OK
¦
¦
¦
¦
¦
¦ ----------------------¦
16.5 Balance History
The balance history function allows you to view and print the history of the balance.
a) Press « » to start "BAL.HIST" .
b) Press « » to scroll forward through the displayed list of balance history information.
c) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOSE".
d) Press « » for printout.
Sample Information Displayed:
Information
Operation Time (year:day:hour)
Total load kg (t)
Number of weighings
Number of key pressed
Number of motor movements
Backlight time (year:day:hour)
Next service due date
¦
¦
¦
¦
¦
¦
Sample Printout:
¦ -- Statistical Info -¦
11:34 ¦ ¦ 21.Jan. 2009
¦
¦ METTLER TOLEDO
¦
¦
Balance Type
SNR:
Balance ID:
SW
MS204S
1234567890
LAB-3
4.23
¦
¦
¦
¦
¦
¦
Display
110:321:23
1485.1345 kg
999999999999
999999999999
9999999999
10:32:23
01:01:2010 dat
63
Application "Diagnostics"
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
Operating hours
10y 321d 23h
Total weight loaded
1485.1345 t
Number of weights
999999999999
Number of key press
999999999999
Motor movements
9999999999
Backlight runnng hours
10y 321d 23h
Next service due date
01.01.2010
¦ ------------------------
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
16.6 Calibration History
The "Calibration History" function allows you to view and print information on the last 30 (thirty) balance adjust ment. Adjustments made by a service technician and normal user are counted together.
a) Press « ».to start "CAL.HIST".
b) Press « » key to scroll forward through the displayed list of Adjustments history information.
c) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOSE".
d) Press « » for printout.
Sample Information Displayed:
Note
S = External adjusted service
I = Internal adjusted
Display
06:03:09 S
05:03:09 I
·
·
·
04:03:09 F
03:03:09 I
02:03:09 E
05:03:09 I
F = FACT
I = Internal adjusted
E = External adjusted user
I = Internal adjusted
Sample Printout:
¦ ------ Calibration -----¦
11:34 ¦ ¦ 05.Mar. 2009
¦
¦ METTLER TOLEDO
¦
¦
¦
¦
¦
Balance Type
SNR:
MS204S
1234567890
¦ --------------------------- ¦
¦
¦
¦
64
Application "Diagnostics"
¦
¦
¦
¦
01 05.Mar. 2009
External
ADJ SERVICE
11:34
100.0000 g
23.5°C
¦
¦
¦
¦ --------------------------- ¦
02 04.Mar. 2009
FACT
09:00
22.4°C
¦
¦
¦
¦
¦
¦
¦
¦ --------------------------- ¦
¦
¦
¦
¦
¦
·
·
·
28 03.Mar. 2009
FACT
10:59
22.6°C
¦ --------------------------- ¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
29 02.Mar. 2009
External
ADJ USER
16:34
100.0000 g
24.6°C
¦
¦
¦
¦
¦
¦
¦
¦ --------------------------- ¦
30 02.Mar. 2009
FACT
18:36
22.4°C
¦
¦
¦
¦ --------------------------- ¦
16.7 Balance Information
The balance information function allows you to view and print information about your balance.
a) Press « » to start "BAL.INFO".
b) Press « » to scroll forward through the displayed list of Balance information.
c) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOSE".
Sample information displayed:
Information
Balance type
Max. load
Software platform
Serial number
Type definition number
Software version
Cell ID
Cell type
Tolerance revision number
Language
Display
TYPE MS6002S
MAX 6200 g
PLATFORM RAINBOW
SNR 1234567890
TDNR 9.6.3.411.203
SOFTWARE V1.00
CELL ID 1172400044
CELL TYPE MMAI6000G2
TOLERANCE NO2
LANGUAGE ENGLISH
65
Application "Diagnostics"
Sample Printout:
¦ -- Balance Information ¦
¦ 21.Jan. 2009
¦
¦ METTLER TOLEDO
11:34 ¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
Balance Type
Max. Load
SNR:
SW:
Platform
TDNR:
Cell ID:
Cell Type
Tol.Rev.no.
Language
MS6002S
6200 g
1234567890
V1.00
Rainbow
9.6.3.411.2-
03
1172400044
MMAI6000G2
2
English
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦ -----------------------¦
16.8 Service Provider Information
The service provider Information function allows you to print information about your service provider.
a) Press « » to start "PROVIDER".
b) Press « ». The service provider information will be printed and the balance will return to the topic
"DIAGNOSE".
Sample Printout:
¦ -- Service Provider -¦
¦ 21.Jan. 2009
¦
¦
¦
¦
¦
¦
¦ METTLER TOLEDO
Im Langacher
CH-8606 Greifensee
Switzerland
(+41) 044 944 22 11
11:34 ¦
¦ ----------------------¦
¦
¦
¦
¦
¦
¦
¦
66
Error and Status Messages
17 Error and Status Messages
17.1 Error Messages
Error messages in the display draw your attention to incorrect operation or that the balance could not execute a procedure properly.
Error Message
NO STABILITY
WRONG ADJUSTMENT WEIGHT
REFERENCE TOO SMALL
Cause
No stability.
Wrong adjustment weight on pan or none at all.
Reference for piece counting too small.
EEPROM (memory) error.
Rectification
Ensure more stable ambient condi tions. If not possible, check settings for environment.
Place required adjustment weight in center of pan.
Increase reference weight.
EEPROM ERROR - PLEASE CONTACT
CUSTOM SERVICE
WRONG CELL DATA - PLEASE CON
TACT CUSTOM SERVICE
NO STANDARD ADJUSTMENT -
PLEASE CONTACT CUSTOM SERVICE
PROGRAM MEMORY DEFECT -
PLEASE CONTACT CUSTOM SERVICE
TEMP SENSOR DEFECT - PLEASE
CONTACT CUSTOM SERVICE
WRONG LOAD CELL BRAND -
PLEASE CONTACT CUSTOMER SER
VICE
WRONG TYPE DATA SET - PLEASE
CONTACT CUSTOMER SERVICE
Wrong cell data.
No standard calibration.
Program memory defect.
Temperature sensor defect.
Wrong load cell brand.
Wrong type data set.
Overload - The weight on the pan exceeds the weighing capacity of the balance.
Underload
Please contact METTLER TOLEDO customer service.
Please contact METTLER TOLEDO customer service.
Please contact METTLER TOLEDO customer service.
Please contact METTLER TOLEDO customer service.
Please contact METTLER TOLEDO customer service.
Please contact METTLER TOLEDO customer service.
Please contact METTLER TOLEDO customer service.
Reduce the weight on the weighing pan.
MEM FULL
FACTOR OUT OF RANGE
STEP OUT OF RANGE
OUT OF RANGE
BELOW ZERO RANGE
Memory full.
Factor is outside the allow range.
Step is outside the allow range.
Sample weight is outside the allow range.
Belong zero range.
Check that the weighing pan is posi tioned correctly.
Clear the memory and start a new evaluation.
Select a new factor.
Select a new step.
Unload the pan and load a new sample weight.
Check that the weighing pan is posi tioned correctly.
17.2 Status Messages
Status messages are displayed by means of small icons. The status icons indicate the following:
67
Error and Status Messages
Status Icoon Signification
Service ReminderYour balance is due for servicing. Contact your dealer's customer service department as soon as possible to have a technician service your balance. (see menu topic "SERV.ICON")
68
Cleaning and Service
18 Cleaning and Service
Every now and then, clean the weighing pan, draft shield element, bottom plate, draft shield (depending on the model) and housing of your balance. Your balance is made from high-quality, durable materials and can therefore be cleaned using a damp cloth or with a standard, mild cleaning agent.
To thoroughly clean the draft shield glass panels, remove the draft shield from the balance. When reinstalling the draft shield, ensure that it is in the correct position.
Please observe the following notes:
●
●
●
The balance must be disconnected from the power supply
Ensure that no liquid comes into contact with the balance or the AC adapter.
Never open the balance or AC adapter – they contain no components, which can be cleaned, repaired or replaced by the user.
● On no account use cleaning agents, which contains solvents or abrasive ingredients, as this can result in damage to the operation panel overlay.
●
Do not clean the IP65 protected models using high-pressure or high-temperature water.
Please contact your METTLER TOLEDO dealer for details of the available service options. Regular servicing by an authorized service engineer ensures constant accuracy for years to come and prolongs the service life of your balance.
18.1 Cleaning the Glass Draft Shield (0.1 mg and 1 mg Models)
Remove the following parts:
a) Remove weighing pan, draft shield element (0.1 mg models) and pan support.
b) Remove the bottom plate.
c) Unlock the draft shield, lift it off the balance and place it on a clean surface.
69
Cleaning and Service
B
C
A
a) Push the glass doors (A) back.
b) Turn the two lock covers (B) on the front as far as it will go.
a) Tilt the front glass (C) forward.
b) Remove the front-glass.
Pull the top glass door (D)out from the front.
F
E
Lift the side glass doors (E) at (F) and pull it out from the front.
70
Cleaning and Service
H
G
Push the lock button (G) to release the rear glass.
Remove the rear glass (H).
a) Turn the draft shield lock to the " " (Service) position.
b) Remove the draft shield lock.
After cleaning reinstall all components in the reverse order. For balance mounting see chapter "Setting up the
Balance – Installing the Components".
71
Interface Specification
19 Interface Specification
19.1 RS232C Interface
Each balance is equipped with an RS232C Interface as standard for the attachment of a peripheral device (e.g.
printer or computer).
Schematic
GND
5
9 6
RxD
TxD
1
CTS
RTS
DATA
IN
OUT
HAN D
SHA KE
IN
OUT
Item Specification
Interface type Voltage interface according to EIA
RS-232C/DIN66020 CCITT V24/V.28)
Max. cable length
15 m
Signal level
Connector
Operating mode
Outputs:
+5 V … +15 V (RL = 3–7 kΩ)
–5 V … –15 V (RL = 3–7 kΩ)
Inputs:
+3 V … +25 V
–3 V … -25 V
Sub-D, 9-pole, female
Full duplex
Bit-serial, asynchronous Transmission mode
Transmission code
Baude rates
ASCII
Bits/parity
Stop bits
Handshake
End-of-line
300, 1200, 2400, 4800, 9600, 19200, 38400
(software selectable)
7-bit/none, 7-bit/even, 7-bit/odd, 8-bit/none (soft ware selectable)
1 stop bit
None, XON/XOFF, RTS/CTS (software selectable)
<CR><LF>, <CR>, <LF> (software selectable)
19.2 USB Device Interface
Each balance is equipped with an "USB Device" Interface as standard for the attachment of a peripheral device
(e.g. computer).
Note: This interface is not suitable to communicate with a Printer.
72
Schematic
Interface Specification
Item Specification
Standard
Speed
In conformity with USB Specification Revision 1.1
Full speed 12 Mbps (requires shielded cable)
Function CDC (Communication Device Class) serial port emulation
Power usage Suspended device: Max 10 mA
Connector Type B
1
2
3
4
VBUS (+5 VDC)
D– (Data –)
D+ (Data +)
GND (Ground)
Shield Shield
19.3 MT-SICS Interface Commands and Functions
Many of the balances and scales used have to be capable of integration in a complex computer or data acquisition system.
To enable you to integrate balances in your system in a simple manner and utilize their capabilities to the full, most balance functions are also available as appropriate commands via the data interface.
All new METTLER TOLEDO balances launched on the market support the standardized command set "METTLER
TOLEDO Standard Interface Command Set" (MT-SICS). The commands available depend on the functionality of the balance.
Basic information on data interchange with the balance
The balance receives commands from the system and acknowledges the command with an appropriate response.
Command formats
Commands sent to the balance comprised one or more characters of the ASCII character set. Here, the following must be noted:
● Enter commands only in uppercase.
●
●
The possible parameters of the command must be separated from one another and from the command name by a space (ASCII 32 dec., in this description represented as ).
The possible input for "text" is a sequence of characters of the 8-bit ASCII character set from 32 dec to 255 dec.
● Each command must be closed by C
R
L
F
(ASCII 13 dec., 10 dec.). The characters C
R
L
F
, which can be inputted using the Enter or Return key of most entry keypads, are not listed in this description, but it is essential they be included for communication with the balance.
Example
S – Send stable weight value
Command
Response
S
S S WeightValue Unit
Get the current stable net weight value.
Current stable weight value in unit actually set under unit 1.
73
Interface Specification
S I
S +
S -
Command not executable (balance icurrently exe cuting another command, e.g. taring, or timeout as stability was not reached).
Balance in overload range.
Balance in underload range.
Example
Command
Response
S
S S 100.00 g
Query a stable weight value.
The current stable weight value is 100.00 g.
The available MT-SICS commands are listed in the table. For further information please refer to the Reference Manu al "MT-SICS 11780711" downloadable from the Internet under www.mt.com/sics-newclassic.
Description Description
@
C0
C1
C2
C3
D
DAT
DW
I0
I1
I2
I3
I4
I5
I10
I11
I14
K
M02
M03
M08
M09
M11
M14
Cancel (Reset)
Query/Set adjustment settings
Start adjustment according to current settings PW
Start adjustment with external weight
Start adjustment with internal weight
Display text sent to balance
Date query/set
Display weigh
Commands implemented
MT-SICS level and MT-SICS versions
Balance data
Software version, type definition number
Query serial number (SNR)
Query SW-identification number
Query balance ID
Query balance type
Query balance information
Keys: set configuration
Query/set environment
Query/set AutoZero
Display brightness
Display contrast
Beeper: Query/set volume
List available language
M31
M46
PWR
S
SI
SIR
SIRU
SIU
SM0
Mode after restart
Print interval
Piece counting: Query/Set piece weight
Power on/off (PWR 0 means switch off com pletely, if balance is powered by battery)
Send stable weight value
Send weight value immediately
Send weight value immediately and repeat
Send weight value with currently displayed unit immediately and repeat
Send weight value with currently displayed unit immediately
Dynamic weighing: cancel all SMx com mands
SM1
SM2
SM3
Dynamic weighing: Start immediately and send the result
Dynamic weighing: start after a minimum load is exceeded and send result
Dynamic weighing: start after a minimum load is exceeded, send result and repeat
SM4
SNR
SNRU
Send stable weight valuewith currently dis played unit and repeat on weight change
SR
SRU
Send weight value on weight change
Send stable weight value with currently dis played unit on weight change
ST
Dynamic weighing: query/set time interval
Send stable weight value and repeat on weight change
SU
Send stable weight value on pressing (print) key
Send stable weight value with currently dis played unit
T
TA
TAC
TI
Tare
Get/Set tare weight value
Clear tare value
Tare immediately
74
M15
M17
M22
M25
M26
M27
M30
Description
Query/set language
FACT: query/set single time criteria (no possi bility to set "weekday", 0 and 127)
Custom unit definition remarks: only one cus tom unit available, no possibility to set
"name" of unit
Get application list
Get/set current application
Adjustment history
+/- settings with nominal and tolerance
TIM
TST0
TST1
TST2
TST3
UPD
Z
ZI
Interface Specification
Description
Query/set time
Query/set test function settings
Start test function according to current set tings
Start test function with external weight
Start test function with internal weight
Query/set update rate of the host interface
Zero
Zero immediately
75
Technical Data
20 Technical Data
20.1 General Data
Power Supply
• S Platform:
• L Platform:
Protection and Standards
• Overvoltage categorie:
• Degree of pollution:
• Degree of Protection:
• Standards for safety and EMC:
• Range of application:
Environmental conditions
• Heigh above mean sea level:
• Ambient temperature range:
• Relative air humidity:
Materials
• Housing:
• Weighing pan:
• Draft shield element:
• Draft shield:
• In-use-cover:
AC/DC Adapter
Primary: 100V–240V, 50/60Hz, 0.3 A
Secondary: 12VDC, 0.84A (with electronic overload protection)
Power supply to the balance: 11–20VDC, 10W
Use only with a tested AC Adapter with SELV output current.
Ensure correct polarity
Power supply 100V–240V, 50/60Hz, 0.3 A
Power cable 2-core with country-specific plug
MS-KL models: Built-in rechargeable NiMH battery (nickel-metal hydride)
Class II
2
Protected against dust and water
Models with S + L Platform: IP54 in use with weighing pan
MS-KLIP models: IP65
See Declaration of Conformity
For use only in enclosed interior rooms up to 4000 m
10 to 30 °C (S platform)
5 to 40 °C (L platform, MS-L models)
-10 to 40 °C (L platform, MS-KL models)
10% to 80 % at 31 °C, linearly decreasing to 50 % at 40 °C, noncon densing
Die-cast aluminum, lacquered
Stainless steel X2CrNiMo 17-12-3 (1.4404)
245 x 351 mm: Stainless steel X5CrNiMo 18-10 (1.4301) with 0.1 mg models: Stainless steel X2CrNiMo 17-12-3 (1.4404) with 10 mg models: Plastic (PBT)
Plastic (PBT), glass
Plastic (PBT)
76
Technical Data
20.2 Model-Specific Data
20.2.1 Balances with Readability of 0.1 mg, S Platform with Draft Shield
Technical Data
Model
Maximum load
Maximum load, fine range
Readability
Readability, fine range
Taring range
Repeatability (sd)
Repeatability (sd), fine range
Linearity
Linearity, fine range
Sensitivity temperature drift (10...30°C)
Eccentric load deviation (test load)
Internal adustment
Adjustment range with external weights
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)
Minimum weight (U=1%, k=2)
Minimum weight (OIML)
Settling time, typ.
Weighing technologiy
Usable heigh of draftshield [mm]
Weighing pan dimensions (WxD) [mm]
Balance dimensions (WxDxH) [mm]
Net Weight [kg]
MS104S
120 g
–
0.1 mg
–
0...120 g
0.1 mg
–
0.2 mg
–
1.5 ppm/°C
0.2 mg (50 g) yes, FACT
50...120 g
100 g / F2/4
5 g / E2/2
0.3 g
0.02 g
0.01 g
2 s
MonoBloc
237
Ø 90
204x347x345
6.5
M204S
220 g
–
0.1 mg
–
0...220 g
0.1 mg
–
0.2 mg
–
1.5 ppm/°C
0.3 mg (100 g) yes, FACT
100...220 g
200 g / F2/4
10 g / F1/3
0.3 g
0.02 g
0.01 g
2 s
MonoBloc
237
Ø 90
204x347x345
6.5
MS304S
320 g
–
0.1 mg
–
0...320 g
0.1 mg
–
0.3 mg
–
1.5 ppm/°C
0.3 mg (200 g) yes, FACT
100...320 g
200 g / F2/4
10 g / F1/3
0.3 g
0.02 g
0.01 g
3 s
MonoBloc
237
Ø 90
204x347x345
6.5
20.2.2 Balances with Readability of 1 mg, S Platform with Draft Shield
Technical Data
Model
Maximum load
Maximum load, fine range
Readability
Readability, fine range
Taring range
Repeatability (sd)
Repeatability (sd), fine range
Linearity
Linearity, fine range
Sensitivity temperature drift (10...30°C)
MS303S
320 g
–
0.001 g
–
0...320 g
0.001 g
–
0.002 g
–
3 ppm/°C
MS303SE
320 g
–
0.001 g
–
0...320 g
0.001 g
–
0.002 g
–
3 ppm/°C
MS403S
420 g
–
0.001 g
–
0...420 g
0.001 g
–
0.002 g
–
3 ppm/°C
77
Technical Data
Model
Eccentric load deviation (test load)
Internal adustment
Adjustment range with external weights
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)
Minimum weight (U=1%, k=2)
Minimum weight (OIML)
Settling time, typ.
Weighing technologiy
Usable heigh of draftshield [mm]
Weighing pan dimensions (WxD) [mm]
Balance dimensions (WxDxH) [mm]
Net Weight [kg]
Model
Maximum load
Maximum load, fine range
Readability
Readability, fine range
Taring range
Repeatability (sd)
Repeatability (sd), fine range
Linearity
Linearity, fine range
Sensitivity temperature drift (10...30°C)
Eccentric load deviation (test load)
Internal adustment
Adjustment range with external weights
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)
Minimum weight (U=1%, k=2)
Minimum weight (OIML)
Settling time, typ.
Weighing technologiy
Usable heigh of draftshield [mm]
Weighing pan dimensions (WxD) [mm]
Balance dimensions (WxDxH) [mm]
Net Weight [kg]
MS303S
3 mg (200 g) yes, FACT
100...320 g
200 g / F2/4
10 g / F1/3
3 g
0.2 g
0.02 g
1.5 s
MonoBloc
165
127x127
204x347x280
6.2
MS603S
620 g
–
0.001 g
–
0...620 g
0.001 g
–
0.002 g
–
3 ppm/°C
3 mg (200 g) yes, FACT
100...620 g
500 g / F2/4
20 g / F1/3
3 g
0.2 g
0.02 g
1.5 s
MonoBloc
165
127x127
204x347x280
6.2
MS303SE
3 mg (200 g) no, EXT ADJ
100...320 g
200 g / F2/4
10 g / F1/3
3 g
0.2 g
0.02 g
1.5 s
MonoBloc
165
127x127
204x347x280
6.2
MS1003S
1020 g
–
0.001 g
–
0...1020 g
0.001 g
–
0.002 g
–
3 ppm/°C
3 mg (500 g) yes, FACT
500...1020 g
1000 g / F2/4
50 g / F2/4
3 g
0.2 g
0.02 g
1.5 s
MonoBloc
165
127x127
204x347x280
6.9
MS403S
3 mg (200 g) yes, FACT
100...420 g
200 g / F2/4
20 g / F1/3
3 g
0.2 g
0.02 g
1.5 s
MonoBloc
165
127x127
204x347x280
6.2
78
Technical Data
20.2.3 Balances with Readability of 0.01 g, S Platform
Technical Data
Model
Maximum load
Maximum load, fine range
Readability
Readability, fine range
Taring range
Repeatability (sd)
Repeatability (sd), fine range
Linearity
Linearity, fine range
Sensitivity temperature drift (10...30°C)
Eccentric load deviation (test load)
Internal adustment
Adjustment range with external weights
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)
Minimum weight (U=1%, k=2)
Minimum weight (OIML)
Settling time, typ.
Weighing technologiy
Weighing pan dimensions (WxD) [mm]
Balance dimensions (WxDxH) [mm]
Net Weight [kg]
Model
Maximum load
Maximum load, fine range
Readability
Readability, fine range
Taring range
Repeatability (sd)
Repeatability (sd), fine range
Linearity
Linearity, fine range
Sensitivity temperature drift (10...30°C)
Eccentric load deviation (test load)
Internal adustment
Adjustment range with external weights
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)
Minimum weight (U=1%, k=2)
MS1602S
1620 g
–
0.01 g
–
0...1620 g
0.01 g
–
0.02 g
–
3 ppm/°C
30 mg (1000 g) yes, FACT
1000...1620 g
1000 g / F2/4
50 g / F2/4
30 g
2 g
0.5 g
1.5 s
MonoBloc
170x200
194x347x97
5.5
MS3002SE
3200 g
–
0.01 g
–
0...3200 g
0.01 g
–
0.02 g
–
3 ppm/°C
30 mg (2000 g) no, EXT ADJ
1000...3200 g
1000 g / F2/4
100 g / F2/4
30 g
2 g
MS1602SE
1620 g
–
0.01 g
–
0...1620 g
0.01 g
–
0.02 g
–
3 ppm/°C
30 mg (1000 g) no, EXT ADJ
1000...1620 g
2000 g / F2/4
200 g / F2/4
30 g
2 g
1000 g / F2/4
50 g / F2/4
30 g
2 g
0.5 g
1.5 s
MonoBloc
170x200
194x347x97
5.5
MS4002S
4200 g
–
0.01 g
–
0...4200 g
0.01 g
–
0.02 g
–
3 ppm/°C
30 mg (2000 g) yes, FACT
1000...4200 g
MS3002S
3200 g
–
0.01 g
–
0...3200 g
0.01 g
–
0.02 g
–
3 ppm/°C
30 mg (2000 g) yes, FACT
1000...3200 g
1000 g / F2/4
100 g / F2/4
30 g
2 g
0.5 g
1.5 s
MonoBloc
170x200
194x347x97
5.5
MS4002SDR
4200 g
820 g
0.1 g
0.01 g
0...4200 g
0.06 g
0.01
0.2 g
0.02 g
3 ppm/°C
30 mg (2000 g) yes, FACT
1000...4200 g
2000 g / F2/4
200 g / F2/4
30 g
2 g
79
Technical Data
Model
Minimum weight (OIML)
Settling time, typ.
Weighing technologiy
Weighing pan dimensions (WxD) [mm]
Balance dimensions (WxDxH) [mm]
Net Weight [kg]
Model
Maximum load
Maximum load, fine range
Readability
Readability, fine range
Taring range
Repeatability (sd)
Repeatability (sd), fine range
Linearity
Linearity, fine range
Sensitivity temperature drift (10...30°C)
Eccentric load deviation (test load)
Internal adustment
Adjustment range with external weights
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)
Minimum weight (U=1%, k=2)
Minimum weight (OIML)
Settling time, typ.
Weighing technologiy
Weighing pan dimensions (WxD) [mm]
Balance dimensions (WxDxH) [mm]
Net Weight [kg]
MS3002SE
0.5 g
1.5 s
MonoBloc
170x200
194x347x97
5.5
MS6002S
6200 g
–
0.01 g
–
0...6200 g
0.01 g
–
0.02 g
–
3 ppm/°C
30 mg (2000 g) yes, FACT
2000...6200 g
5000 g / F2/4
200 g / F2/4
30 g
2 g
0.5 g
1.2 s
MonoBloc
170x200
194x347x97
5.3
20.2.4 Balances with Readability of 0.1 g, S Platform
Technical Data
Model
Maximum load
Maximum load, fine range
Readability
Readability, fine range
Taring range
Repeatability (sd)
Repeatability (sd), fine range
Linearity
Linearity, fine range
MS6001S
6200 g
–
0.1 g
–
0...6200 g
0.1 g
–
0.2 g
–
MS8001S
8200 g
–
0.1 g
–
0...8200 g
0.1 g
–
0.2 g
–
MS4002S
0.5 g
1.5 s
MonoBloc
170x200
194x347x97
5.5
MS6002SDR
6200 g
1200 g
0.1 g
0.01 g
0...6200 g
0.06 g
0.01 g
0.2 g
0.02 g
3 ppm/°C
30 mg (2000 g) yes, FACT
2000...6200 g
5000 g / F2/4
200 g / F2/4
30 g
2 g
0.5 g
1.2 s
MonoBloc
170x200
194x347x97
5.3
MS4002SDR
0.5 g
1.5 s
MonoBloc
170x200
194x347x97
5.5
MS8001SE
8200 g
–
0.1 g
–
0...8200 g
0.1 g
–
0.2 g
–
80
Model
Sensitivity temperature drift (10...30°C)
Eccentric load deviation (test load)
Internal adustment
Adjustment range with external weights
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)
Minimum weight (U=1%, k=2)
Minimum weight (OIML)
Settling time, typ.
Weighing technologiy
Weighing pan dimensions (WxD) [mm]
Balance dimensions (WxDxH) [mm]
Net Weight [kg]
MS6001S
5 ppm/°C
0.2 g (2000 g) yes, FACT
2000...6200 g
5000 g / F2/4
200 g / F2/4
300 g
20 g
5 g
1 s
MonoBloc
190x226
194x347x97
5.6
MS8001S
5 ppm/°C
0.2 g (5000 g) yes, FACT
2000...8200 g
5000 g / F2/4
200 g / F2/4
300 g
20 g
5 g
1 s
MonoBloc
190x226
194x347x97
5.6
20.2.5 Balances with Readability of 0.1 g to 1 g, L Platform
Technical Data
Model
Maximum load
Maximum load, fine range
Readability
Readability, fine range
Taring range
Repeatability (sd)
Repeatability (sd), fine range
Linearity
Linearity, fine range
Sensitivity temperature drift (10...30°C)
Eccentric load deviation (test load)
Internal adustment
Adjustment range with external weights
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)
Minimum weight (U=1%, k=2)
Minimum weight (OIML)
Settling time, typ.
Weighing technologiy
Built-in battery
Weighing below the balance (with optional hook)
Weighing pan dimensions (WxD) [mm]
Balance dimensions (WxDxH) [mm]
Net Weight [kg]
MS12001L
12200 g
–
0.1 g
–
0...12200 g
0.1 g
–
0.2 g
–
5 ppm/°C
0.3 g (5000 g) yes, FACT
5000...12200 g
10000 g / F2/4
500 g / F2/4
300 g
20 g
5 g
2 s
MonoBloc no yes
351x245
363x346x118
10.7
MS16001L
16200 g
–
0.1 g
–
0...16200 g
0.1 g
–
0.2 g
–
5 ppm/°C
0.3 g (5000 g) yes, FACT
5000...16200 g
10000 g / F2/4
500 g / F2/4
300 g
20 g
5 g
2 s
MonoBloc no yes
351x245
363x346x118
10.7
MS16001LE
16200 g
–
0.1 g
–
0...16200 g
0.1 g
–
0.2 g
–
15 ppm/°C
0.3 g (5000 g) no, EXT ADJ
5000...16200 g
10000 g / F2/4
500 g / F2/4
300 g
20 g
5 g
2 s
MonoBloc no yes
351x245
363x346x118
10.7
Technical Data
MS8001SE
15 ppm/°C
0.2 g (5000 g) no, EXT ADJ
2000...8200 g
5000 g / F2/4
200 g / F2/4
300 g
20 g
5 g
1 s
MonoBloc
190x226
194x347x97
5.6
81
Technical Data
Model
Maximum load
Maximum load, fine range
Readability
Readability, fine range
Taring range
Repeatability (sd)
Repeatability (sd), fine range
Linearity
Linearity, fine range
Sensitivity temperature drift (10...30°C)
Eccentric load deviation (test load)
Internal adustment
Adjustment range with external weights
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)
Minimum weight (U=1%, k=2)
Minimum weight (OIML)
Settling time, typ.
Weighing technologiy
Built-in battery
Weighing below the balance (with optional hook)
Weighing pan dimensions (WxD) [mm]
Balance dimensions (WxDxH) [mm]
Net Weight [kg]
MS32001L MS32001LE
32200 g
–
0.1 g
–
0...32200 g
0.1 g
–
0.3 g
–
32200 g
–
0.1 g
–
0...32200 g
0.1 g
–
0.3 g
–
5 ppm/°C
0.3 g (5000 g) yes, FACT
15 ppm/°C
0.3 g (5000 g) no, EXT ADJ
10000...32200 g 10000...32200 g
10000 g / F2/4
500 g / F2/4
300 g
20 g
5 g
1.5 s
MonoBloc no yes
351x245
363x346x118
10.7
10000 g / F2/4
500 g / F2/4
300 g
20 g
5 g
1.5 s
MonoBloc no yes
351x245
363x346x118
10.7
Model
Maximum load
Maximum load, fine range
Readability
Readability, fine range
Taring range
Repeatability (sd)
Repeatability (sd), fine range
Linearity
Linearity, fine range
Sensitivity temperature drift (10...30°C)
Eccentric load deviation (test load)
Internal adustment
Adjustment range with external weights
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)
Minimum weight (U=1%, k=2)
MS32000L MS32000LE
32200 g
–
1 g
–
0...32200 g
0.5 g
–
32200 g
–
1 g
–
0...32200 g
0.5 g
–
1 g
–
5 ppm/°C
1 g
–
15 ppm/°C
1 g (10000 g) yes, FACT
1 g (10000 g) no, EXT ADJ
10000...32200 g 10000...32200 g
20000 g / F2/4
1000 g / F2/4
1500 g
100 g
20000 g / F2/4
1000 g / F2/4
1500 g
100g
82
Model
Minimum weight (OIML)
Settling time, typ.
Weighing technologiy
Built-in battery
Weighing below the balance (with optional hook)
Weighing pan dimensions (WxD) [mm]
Balance dimensions (WxDxH) [mm]
Net Weight [kg]
MS32000L
50 g
1.5 s
MonoBloc no yes
351x245
363x346x118
10.7
MS32000LE
50 g
1.5 s
MonoBloc no yes
351x245
363x346x118
10.6
20.2.6 Balances with Readability of 2 g to 5 g, L Platform
Technical Data
Model
Maximum load
Readability
Weighing range(approved version)
Readability(approved version)
Taring range
Repeatability (sd)
Linearity
Internal adustment
Adjustment range with external weights
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Settling time, typ.
Weighing technologiy
Weighing pan dimensions (WxD) [mm]
Built-in battery
IP Protection
Weighing below the balance (with optional hook)
Balance dimensions (WxDxH) [mm]
Net Weight [kg]
MS15KLE
15 kg
2 g
6 kg / 15 kg
2 g / 5 g
0...15 kg
1 g
2 g no, EXT ADJT
5...15 kg
10 kg / F2/4
500 g / F2/4
1 s
Strain Gauge
351x245 yes
IP 54 in use no
363x346x122
9.9
Model
Maximum load
Readability
Weighing range(approved version)
Readability(approved version)
Taring range
Repeatability (sd)
Linearity
Internal adustment
Adjustment range with external weights
MS24KLIPE
24 kg
2 g
15 kg / 24 kg
5 g / 10 g
0...24 kg
2 g
2 g no, EXT ADJT
10...24 kg
MS15KLIPE
15 kg
2 g
6 kg / 15 kg
2 g / 5 g
0...15 kg
1 g
2 g no, EXT ADJ
5...15 kg
10 kg / F2/4
500 g / F2/4
1 s
Strain Gauge
351x245 yes
IP 65 no
363x346x122
9.9
MS30KLE
30 kg
2 g
15 kg / 30 kg
5 g / 10 g
0...30 kg
2 g
2 g no, EXT ADJ
10...30 kg
Technical Data
83
Technical Data
Model
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Settling time, typ.
Weighing technologiy
Weighing pan dimensions (WxD) [mm]
Built-in battery
IP Protection
Weighing below the balance (with optional hook)
Balance dimensions (WxDxH) [mm]
Net Weight [kg]
MS24KLIPE
20 kg / F2/4
1000 g / F2/4
1 s
Strain Gauge
351x245 yes
IP 65 no
363x346x122
9.9
MS30KLE
20 kg / F2/4
1000 g / F2/4
1 s
Strain Gauge
351x245 yes
IP 54 in use no
363x346x122
9.9
84
Technical Data
20.3 Dimensions
20.3.1 Balances with Readability of 0.1 mg, S Platform With Draft Shield
Models:
MS104S
MS204S
MS304S
85
ø 90
86
201
246
218.5
21
346.5
173
85
Technical Data
20.3.2 Balances with Readability of 1 mg, S Platform With Draft Shield
Models:
MS303S
MS303SE
MS403S
MS603S
MS1003S
85
86
104.5
10.5
246
218.5
21
346.5
127
173
86
20.3.3 Balances with Readability of 0.01 g, S Platform
Models:
MS1602S
MS1602SE
MS3002S
MS3002SE
MS4002S
MS4002SDR
MS6002S
MS6002SDR 104.5
85.5
246
28
19.5
Technical Data
200
110
346.5
236.5
87
Technical Data
20.3.4 Balances with Readability of 0.1 g, S Platform
Models:
MS6001S
MS8001S
MS8001SE
104.5
85.5
246
28
14
226
106
346.5
240.5
88
20.3.5 Balances with Readability of 0.1 g to 5 g, L Platform
Models:
MS12001L
MS16001L
MS16001LE
MS32001L
MS32001LE
MS30000LE
MS32000LE
MS15KLE
MS15KLIPE
MS24KLIPE
MS30KLE
MSxxxL
MSxxxKL
40
1)
91.5
230
Technical Data
10
245
90
346
256
89
Accessories and Spare Parts
21 Accessories and Spare Parts
21.1 Accessories
Description
Draft Shields
Draft shield with sliding doors "mg" (165 mm)
Draft shield with sliding doors "0.1 mg" (237 mm)
Part No.
12122405
12122404
Printers
RS-P25 printer with RS232C connection to balance 11124300
RS-P26 printer with RS232C connection to balance (with date and time)
RS-P28 printer with RS232C connection to balance (with date, time and applications
11124303
11124304
Cables for RS232C Interface
RS232 - USB converter – intelligent expansion module for con nection to PC
11103691
90
RS9 – RS25 (m/f): connection cable for PC, length = 2 m
Accessories and Spare Parts
11101052
RS9 – RS9 (m/m): connection cable for devices with DB9 (f) socket, length = 1 m
RS232 - USB converter – intelligent expansion module for con nection to PC
Auxiliary Displays
LC/RS-BLD auxiliary display on bench stand with backlit (incl.
RS cable and separate AC adapter)
RS232 auxiliary display
METTLER TOLEDO
000000
AC Adapters
AC adapter universal (EU, USA, AU, UK) 100–240 VAC,
50/60HZ, 0.3 A, 12 V 0.84 A
Protective Covers
Protective cover for models MS-S with draft shield
21250066
11103691
00224200
12120057
11120270
12121850
91
Accessories and Spare Parts
Protective cover for models MS-S without draft shield
Protective cover for models MS-L up to "1 g"
Protective cover for models MS-L "2–5 g"
Anti-theft Devices
Steel cable
Software
LabX direct balance (simple data transfer)
Weighing Below the Balance
Hook for Platform L
Transport Cases
Transport case for S platform balances
11101052
11132565
11124245
12121851
12121852
12121853
11600361
92
Accessories and Spare Parts
Adjustment Weights
OIML / ASTM Weights (with calibration certificate) see www.mt.com/weights
21.2 Spare Parts
Draft Shield
Drawing
1
2
3
6
7
4
5
Pos Description
5 Draft shield lock
6 Botom plate
Draft Shield "165 mm"
1 Top glass with handle
2 Rear glass low
3 Side glass door left low with handle
4 Side glass door right with handle
5 Front glass low
Draft Shield "237 mm"
1 Top glass with handle
2 Rear glass high
3 Side glass door left high with handle
4 Side glass door right high with handle
5 Front glass high
Part No.
12122013
12122019
12121884
12122015
12121881
12121883
12122014
12121884
12122012
12121880
12121882
12122011
93
Accessories and Spare Parts
Weighing Pans / Draft Shield Elements / Support
Drawing Pos Description
C
C
1
2
3
4
5
6
7
8
9
10
11
10
For S platform
1
2
3
4
5
0.1 mg
0.1 mg
0.1 mg
1 mg
1 mg
Weighing pan Ø 90 mm
Pan support
Draft shield element
Weighing pan 127x127 mm
Pan support for models up to 999 g
5 1 mg
6 10 mg
7
7
10 mg
0.1 g
Weighing pan 170x200 mm
Weighing pan 190x226 mm
8 from 10 mg Pan support caps
9 Leveling foot
For L platform
Pan support for models from 1000 g
Draft shield element
10 Weighing pan 246x352 mm
11 to 1 g Pan support caps
12 from 2 g Pan support caps
13 Leveling foot
12
Part No.
12122010
12122007
12122008
12122009
12122017
12122016
12122018
12122003
12122004
12122005
12122002
12122020
12122001
12122006
12122000
13
13
94
Appendix
22 Appendix
22.1 Conversion Table for Weight Units
Kilogram
Milligram
Microgram
Carat
Pound
Ounce (avdp)
Ounce (troy)
1 kg = 1000.0
1 mg = 0.001
1 µg = 0.000001
1 ct = 0.2
1 lb = 453.59237
1 oz = 28.349523125
1 ozt = 31.1034768
Grain
Pennyweight
Momme
1 GN = 0.06479891
1 dwt = 1.55517384
1 mom = 3.75
Mesghal 1 msg ≈ 4.6083
Tael Hong Kong 1 tlh = 37.429
1 tls ≈ 37.7993641666667
g Tael Singapore
(Malaysia)
Tael Taiwan
Tola
1 tlt
1 tola
= 37.5
= 11.6638038
g g g g
Baht 1 baht = 15.16
g g g g g g g g g g g
1 g = 0.001
1 g = 1000.0
kg mg
1 g = 1000000.0
1 g = 5.0
µg ct
1 g ≈ 0.00220462262184878
lb
1 g ≈ 0.0352739619495804
1 g ≈ 0.0321507465686280
oz ozt
1 g ≈ 15.4323583529414
1 g ≈ 0.643014931372560
1 g ≈ 0.266666666666667
1 g ≈ 0.217
1 g ≈ 0.0267172513291833
1 g ≈ 0.0264554714621853
GN dwt mom msg tlh tls
1 g ≈ 0.0266666666666667
1 g ≈ 0.0857353241830079
1 g ≈ 0.0659630606860158
tlt tola baht
95
Index
23 Index
B
Backlight
Balance history
Balance information
Basic menu
Battery operation
Baudrate
Beep
Bit/Parity
C
Calibration
Calibration history
Cancel
Change settings
Char Set
Check Weighing
Cleaning
Closing the menu
Control Limit
A
Accessories
Adjustment
Advanced menu
Advanced Menu
Ambient conditions
Appendix
Application “Check Weighing”
Application “Diagnostics”
Application “Dynamic Weighing”
Application “Formulation”
Application “Multiplication Factor Weighing”
Application “Percent Weighing”
Application “Piece Counting”
Application “Routine Test”
Application “Statistics”
Application “Totaling”
Application “Weighing”
Application icons
Assign Application
Automatic adjustment
Automatic shutoff
Automatic zero setting
Autozero
96
Conventions and symbols
Conversion table for weight units
D
Date
Date format
Delivery inspection
DeltaRange balances
Diagnose
Diagnostics
Diagnostics application
Dimensions
Display
Display panel
Display test
Disposal
Draft Shield
Dynamic Weighing
E
End of Line
Environment
Error messages
External weight
F
FACT
Factor Weighing
Formulation
Fully automatic adjustment
G
Good Weighing Practice
GWP
H
Handshake
I
Icons
Input priciple
Installing the components
Interface menu
Interface MT-SICS commands and functions
Interface RS232C
Interface USB device
Internal weight
Interval
Introduction
Index
97
Index
K
Key assign
Key beep
Key functions
Key test
L
L platform overview
Language
Leveling the balance
Location
M
Manual adjustment with external weight
Manual adjustment with internal weight
Menu
Menu Advanced
Menu Basic
Menu Interface
Menu mode
Menu operation
Menu topic
Motor test
MT-SICS interface commands and functions
Multiplication Factor Weighing
N
Net
Numerical values
O
Opening the menu
Operation keys
Overview
P
Percent Weighing
Performing a simple weighing
Piece Counting
Power supply
Protocol trigger
Q
Quickstart
R
Recall
Repeatability test
98
Reset
Routine Test
RS232C interface
S
S platform overview
Safety precautions
Saving settings
Select menu topic
Selecting the location
Service
Service icon
Service provider information
Service reminder
Set menu mode
Setting up the balance
Shutoff
SOP
Spare parts
Stability beep
Startup
Statistics
Status icons
Status messages
Submenu
Switching the balance on and off
Switching weight units
Symbols and conventions
T
Taring
Technical data dimensions
Technical data general
Technical data model-specific
Time
Time format
Topic
Totaling
Transmit data
Transporting the balance
U
Unit
Unpacking
USB device interface
W
Warning Limit
99
Index
Index
Weighing below the balance
Weighing made simple
Weighing-in aid
Weight unit
Z
Zero range
Zeroing
100
To protect your METTLER TOLEDO product’s future:
METTLER TOLEDO Service assures the quality, measuring accuracy and preservation of value of all METTLER TOLEDO products for years to come.
Please send for full details about our attractive terms of service.
Thank you.
*11781259*
Mettler-Toledo AG
Laboratory & Weighing Technologies
CH-8606 Greifensee, Switzerland
Subject to technical changes
© Mettler-Toledo AG 2009
11781259 0309/2.12
www.mt.com/newclassic
For more information
GWP
®
Good Weighing Practice™
The global weighing guideline GWP ® reduces risks associated with your weighing processes and helps to
• choose the appropriate balance
• reduce costs by optimizing testing procedures
• comply with the most common regulatory requirements
www.mt.com/GWP
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement