Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
Lesen Sie die Bedienungsanleitung vor dem Gebrauch des Gerätes gründlich durch. Beachten Sie insbesondere die Hinweise unter Gliederungspunkt 2. Andernfalls könnten Gesundheits- oder
Sachschäden auftreten. Die Bühler Technologies GmbH haftet nicht bei eigenmächtigen
Änderungen des Gerätes oder für unsachgemäßen Gebrauch.
Read this instruction carefully prior to installation and/or use. Pay attention particularly to all advice and safety instructions to prevent injuries. Bühler Technologies GmbH can not be held responsible for misusing the product or unreliable function due to unauthorised modifications.
BX140013, 02/2011 Art. Nr. 90 32 025
Bühler Technologies GmbH, Harkortstr. 29, D-40880 Ratingen
Tel. +49 (0) 21 02 / 49 89-0, Fax. +49 (0) 21 02 / 49 89-20
Internet: www.buehler-technologies.com
E-mail: [email protected]
1
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
Inhaltsverzeichnis Seite
2
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
Contents Page
BX140013, 02/2011 Art. Nr. 90 32 025
3
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
A
L
T
E
R
N
1 Einleitung
Die Thermotronik TT77 dient zur Überwachung der Temperatur in Fluidsystemen. Es stehen bis zu vier
Schaltausgänge oder ein Schaltausgang in Kombination mit einem Analogausgang zur Verfügung. Die
Anzeige- und Steuereinheit kann entweder zusammen mit dem Schaltrohr auf dem Tank (TT77F...) oder als Fernanzeige (TT77W...) auf einer 35 mm Hutschiene montiert werden. Bei der direkten Montage ist das
Display um 270° schwenkbar.
Die Schaltcharakteristik der Ausgänge kann als Fenster, Hysterese oder als Frequenzausgang gewählt werden. Die Programmierung erfolgt über ein Menü, das an das VDMA Einheitsblatt 24574 ff. angelehnt ist. Eine Logbuchfunktion zur Speicherung der minimalen und maximalen Temperaturwerte sowie von 6
Ereignissen steht zur Verfügung.
Die Konfiguration Ihres Gerätes entnehmen Sie bitte dem Typenschild. Hier finden Sie neben der
Auftragsnummer auch die Artikelnummer und die Typenbezeichnung.
TT77 mit integrierter Anzeige
TT-77F-
Ausführung
MS
VA
Messing
Edelstahl
-2T
-4T
-1T-KT
-2T-KT
-4T-KT
2 x PNP Ausgang
4 x PNP Ausgang
1 x PNP Ausgang /
2 x PNP Ausgang /
4 x PNP Ausgang /
1 x Analogausgang
1 x Analogausgang
1 x Analogausgang
Länge
280
(max. 1000 mm)
370
500
Variabel (bitte angeben)
TT77 mit Fernanzeige
TT-77W-
A
L
T
E
R
N
Ausführung
MS
Messing
VA
Edelstahl
-2T
-4T
-1T-KT
-2T-KT
-4T-KT
2 x PNP Ausgang
4 x PNP Ausgang
1 x PNP Ausgang /
2 x PNP Ausgang /
1 x Analogausgang
1 x Analogausgang
4 x PNP Ausgang / 1 x Analogausgang
Länge
280
(max. 1000mm)
370
500
Variabel (bitte angeben)
1.1 Einsatzbereich
WARNUNG
Alle Gerätetypen sind ausschließlich für industrielle Anwendungen vorgesehen. Es handelt sich nicht um Sicherheitsbauteile. Die Geräte dürfen nicht eingesetzt werden, wenn bei ihrem Ausfall oder bei Fehlfunktion die Sicherheit und Gesundheit von Personen beeinträchtigt wird.
Der Einsatz in explosionsgefährdeten Bereichen ist nicht gestattet.
4
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
2 Wichtige Hinweise
Bitte überprüfen Sie vor Einbau des Gerätes, ob die genannten technischen Daten den Anwendungsparametern entsprechen. Überprüfen Sie ebenfalls, ob alle zum Lieferumfang gehörenden Teile vollständig vorhanden sind.
Der Einsatz der Geräte ist nur zulässig, wenn:
-
das Produkt unter den in der Bedienungs- und Installationsanleitung beschriebenen Bedingungen, dem
Einsatz gemäß Typenschild und für Anwendungen, für die es vorgesehen ist, verwendet wird. Bei eigenmächtigen Änderungen des Gerätes ist die Haftung durch die Bühler Technologies GmbH ausgeschlossen.
-
die im Datenblatt und der Anleitung angegebenen Grenzwerte eingehalten werden.
-
Überwachungsvorrichtung / Schutzvorrichtung korrekt angeschlossen sind.
-
die Service- und Reparaturarbeiten, die nicht in dieser Anleitung beschrieben sind, von Bühler
Technologies GmbH durchgeführt werden.
-
Originalersatzteile verwendet werden.
Diese Bedienungsanleitung ist Teil des Betriebsmittels. Der Hersteller behält sich das Recht vor, die
Leistungs-, die Spezifikations- oder die Auslegungsdaten ohne Vorankündigung zu ändern. Bewahren Sie die Anleitung für den späteren Gebrauch auf.
In dieser Anleitung werden folgende Warnzeichen und Signalwörter benutzt:
Warnung vor einer allgemeinen Gefahr Netzstecker ziehen
Warnung vor dem Einatmen giftiger Gase Atemschutz tragen
Warnung vor ätzenden Flüssigkeiten Gesichtsschutz tragen
Warnung vor explosionsgefährdeten Bereichen
Signalwörter für Warnhinweise:
HINWEIS
Handschuhe tragen
Signalwort für wichtige Information zum Produkt, auf die im besonderen Maße aufmerksam gemacht werden soll.
VORSICHT
Signalwort zur Kennzeichnung einer Gefährdung mit geringem Risiko, die zu einem Sachschaden oder leichten bis mittelschweren Körperverletzungen führen kann, wenn sie nicht vermieden wird.
WARNUNG
Signalwort zur Kennzeichnung einer Gefährdung mit mittlerem Risiko, die möglicherweise
Tod oder schwere Körperverletzungen zur Folge hat, wenn sie nicht vermieden wird.
GEFAHR
Signalwort zur Kennzeichnung einer Gefährdung mit hohem Risiko, die unmittelbar Tod oder schwere Körperverletzung zur Folge hat, wenn sie nicht vermieden wird.
Das Gerät darf nur von Fachpersonal installiert werden, das mit den Sicherheitsanforderungen und den
Risiken vertraut ist.
Beachten Sie unbedingt die für den Einbauort relevanten Sicherheitsvorschriften und allgemein gültigen
Regeln der Technik. Beugen Sie Störungen vor und vermeiden Sie dadurch Personen- und Sachschäden.
BX140013, 02/2011 Art. Nr. 90 32 025
5
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
Der für die Anlage Verantwortliche muss sicherstellen, dass:
-
Sicherheitshinweise und Betriebsanleitungen verfügbar sind und eingehalten werden,
-
Unfallverhütungsvorschriften der Berufsgenossenschaften beachtet werden; in Deutschland:
BGV A1: Grundsätze der Prävention und BGV A3: Elektrische Anlagen und Betriebsmittel,
-
die zulässigen Daten und Einsatzbedingungen eingehalten werden,
-
Schutzeinrichtungen verwendet werden und vorgeschriebene Wartungsarbeiten durchgeführt werden,
-
bei der Entsorgung die gesetzlichen Regelungen beachtet werden.
Wartung, Reparatur:
-
Reparaturen an den Betriebsmitteln dürfen nur von Bühler autorisiertem Personal ausgeführt werden.
-
Nur Umbau-, Wartungs- oder Montagearbeiten ausführen, die in dieser Bedienungs- und
Installationsanleitung beschrieben sind.
-
Nur Original-Ersatzteile verwenden.
-
Bei Durchführung von Wartungsarbeiten jeglicher Art müssen die relevanten Sicherheits- und
Betriebsbestimmungen beachtet werden.
GEFAHR
Giftige, ätzende Gase / Flüssigkeiten
Schützen Sie sich bei allen Arbeiten vor giftigen, ätzenden Gasen /
Flüssigkeiten. Tragen Sie die entsprechende Schutzausrüstung.
3 Aufbauen und Anschließen
Die Montage und der Anschluss dürfen nur von entsprechend ausgebildeten Fachleuten ausgeführt werden. Die jeweils geltenden Sicherheitsvorschriften des Einsatzortes sind einzuhalten!
GEFAHR
Giftige, ätzende Gase / Flüssigkeiten
Schützen Sie sich bei allen Arbeiten vor giftigen, ätzenden Gasen /
Flüssigkeiten. Tragen Sie die entsprechende Schutzausrüstung.
Zum direkten Tankaufbau wird das Schaltrohr in die dafür vorgesehene Bohrung am Tank eingeschraubt.
Bei Installation mit Fernanzeige wird die Anzeige- und Steuereinheit auf einer 35 mm Hutschiene montiert.
Der Sensor wird mit einem Kabel (max. Länge 10 m) verbunden. Der Anschlussstecker für den Sensor befindet sich auf der Unterseite der Anzeige- und Steuereinheit. Spannungsversorgung und Ausgabe der
Signale erfolgen über die Stecker auf der Oberseite.
Das Gerät wird mit einer Nennspannung von 24 V DC versorgt. Die Steckerbelegungen entnehmen Sie bitte der Zeichnung im Anhang.
6
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
Die Schaltausgänge sind als PNP-Transistor ausgeführt (siehe Abbildung).
Hinweis: Bei Messungen des Schaltausgangs mit hochohmigen Messgeräteeingängen oder bei Nutzung als Frequenzausgang ist ein 10 k
Ω Widerstand zwischen Ausgang und Masse (GND) zu schalten, um Fehlmessungen vorzubeugen.
24 V
+
-
F000588x
4 Betrieb
4.1 Einschalten
Wenn im laufenden Betrieb eine Fehlermeldung im Display erscheint, betrachten Sie bitte Kapitel 5.2
„Fehlerbehebung“.
Nach dem das Gerät an die Versorgungsspannung angeschlossen wurde, erscheint zu Beginn kurzeitig die Softwareversion. Direkt im Anschluss wechselt das Display zur Messwertanzeige.
Im Folgenden wird die Funktion der Anzeige- und Steuereinheit beschrieben.
LEDs
Display
Taste ٛ
Taste
ٛ
F000558X
Taste ٛ
4.2 LED-Statusanzeigen
Leuchtdioden oberhalb der Messwertanzeige signalisieren den Status der Temperatur-Schaltausgänge.
Die LEDs sind den Schaltausgängen fest zugeordnet. Die folgende Tabelle zeigt die Werkseinstellungen.
LED 1 – gelb
Status Schaltausgang 1
LED 2 - rot
Status Schaltausgang 2 (falls vorhanden)
LED 3 - gelb
Status Schaltausgang 3 (falls vorhanden)
LED 4 - rot
Status Schaltausgang 4 (falls vorhanden)
Das Schaltverhalten der LED (Leuchten bei geschlossenem oder geöffnetem Schaltkontakt) kann geändert
werden, siehe dazu Kapitel 4.7.7.
BX140013, 02/2011 Art. Nr. 90 32 025
7
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.3 Allgemeine Tastenfunktionen
Die Bedienung erfolgt über die Tasten unterhalb des Displays.
Eine ausführliche Erklärung der Menüsteuerung finden Sie in den folgenden Kapiteln.
Taste Modus: Funktion
- Messwertanzeige: Wechsel der angezeigten Einheiten; Beispiel:
- Im Menü:
- Am Ende des Menüs:
( )
Wechsel in ein untergeordnetes Menü.
Wechsel ins übergeordnete Menü
+ *
(Exit) signalisiert das Ende des Menüs
- Nach Eingabe / Auswahl: Einen eingegebenen Zahlenwert oder eine
Funktionsauswahl bestätigen und speichern.
- Messwertanzeige:
- Im Menü:
- Messwertanzeige:
- Im Menü:
- Im Menü:
Anzeige der Konfiguration
Menüpunkt, Zahlenwert oder Funktionsauswahl aufwärts blättern. Bei gedrückter Taste erfolgt dies fortlaufend.
Wechsel ins Hauptmenü
Menüpunkt, Zahlenwert oder Funktionsauswahl abwärts blättern. Bei gedrückter Taste erfolgt dies fortlaufend.
Verlassen des Haupt- / Unter-/ Wahlmenüs und
Rücksprung zur Messwertanzeige
- Im Menü: Wechsel zur nächst höheren Menüebene
+ *
60 s keine Aktion * - Im Menü:
Verlassen des Haupt- / Unter-/ Wahlmenüs
* Das Verlassen des Wahl- bzw. Einstellmenüs erfolgt ohne Speicherung der geänderten Werte.
Zur Auswahl eines Menüpunktes und zur Einstellung der Werte gehen Sie wie folgt vor:
¾ Öffnen Sie das Hauptmenü mit der Taste .
¾ Wählen Sie das Untermenü mit den Tasten und aus und öffnen Sie das Untermenü mit der
Taste .
¾ Wählen Sie ggf. nächste Untermenü mit den Tasten und und öffnen Sie es mit der Taste .
¾ Wählen Sie den gewünschten Menüpunkt mit den Tasten und aus und öffnen Sie die
Werteliste mit der Taste .
¾ Stellen Sie den Wert mit den Tasten und ein und bestätigen Sie mit der Taste . Die geänderten Einstellungen werden gespeichert und das Gerät kehrt zum Untermenü zurück.
¾ Verlassen Sie das Untermenü, in dem Sie den Menüpunkt EXIT anwählen und mit der Taste bestätigen. Das Gerät kehrt zum übergeordneten Menü bzw. zur Messwertanzeige zurück.
8
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.4 Aktive Tastensperre
Wenn die Tastensperre aktiviert ist, erscheint beim Aufrufen des Menüs mit der Taste die Anzeige
an Stelle des Hauptmenüs. Die aktive Ziffer wird durch einen Punkt gekennzeichnet.
¾
Geben Sie mit den Tasten und den Code ein und bestätigen Sie mit der Taste . Die aktive
Ziffer rückt um eine Stelle nach rechts. Nach Eingabe der 3. Ziffer öffnet sich das Hauptmenü.
Bei einer falschen Eingabe des Zifferncodes springt das Gerät zur Messwertanzeige zurück. Falls Sie das
Passwort vergessen haben, gelangen Sie jederzeit mit dem Mastercode 287 ins Menü.
Sie können die Tastensperre aufheben, in dem Sie im Menüpunkt Loc im Untermenü Basiseinstellungen
Erweiterte Funktionen
b.EF den Code mit der Eingabe 000 zurücksetzen.
4.5 Menü-Übersicht
Die Struktur des Menüs ist angelehnt an das Einheitsblatt 24574 ff. des VDMA. Das Menü ist hierarchisch aufgebaut. In der obersten Menüebene befinden sich die Hauptmenüeinträge, z.B., tEMP , b.EF
, diA , E .
Jedes Hauptmenü enthält weitere Untermenüpunkte.
Die Menüpunkte können je nach Konfiguration des Gerätes variieren. Nicht alle im Nachfolgenden beschriebenen Menüpunkte müssen auf Ihr Gerät zutreffen. Die Konfiguration können Sie abrufen, wenn
Sie im Anzeigemodus die Taste drücken. Angezeigt wird ein 4-stelliger Code, z. B. t s a
Hierbei bedeuten die 4 Stellen tsav: v
F000583x t: Typ t Temperaturmessung o = Niveau- und Temperaturmessung s: Anzahl der Schaltausgänge 1, 2 oder 4 a: Anzahl der Analogausgänge 0 oder 1 v: Montagetyp der Gerätes i = Standardmontage (Tankeinbau)
F
Die einzelnen Menüpunkte erscheinen nicht, wenn die Option nicht vorhanden ist. Beispiel: Bei a=0 sind die Menüpunkte zur Einstellung des Analogausgangs nicht vorhanden. Sie können die Beschreibung dieser Punkte dann überspringen.
Die Struktur der Menüs für die Schalt- und
Analogausgänge ist gleichartig. Hier werden die
Einstellungen für die Schaltausgänge bzw. die
Analogausgänge (falls vorhanden) vorgenommen.
Die Grundeinstellungen des Gerätes können geändert werden. Allgemeingültige Einstellungen erfolgen im Menü Basiseinstellungen Erweiterte
Funktionen
( b.EF
). Diese Einstellungen sollten zuerst vorgenommen werden, da sie sich auf die
Anzeigen und Einstellmöglichkeiten in den einzelnen
Menüs auswirken. Solche Einstellungen sind z. B. die verwendeten Einheiten und die Zuordnung der
Schaltausgänge zu Füllstands- und Temperaturmessung. Die Zuordnung der Analogausgänge kann nicht verändert werden.
Zusätzlich stehen im Menü Diagnostic Möglichkeiten zur Diagnose zur Verfügung.
Die ausführliche Darstellung der gesamten Menüstruktur befindet sich am Ende dieser Anleitung.
BX140013, 02/2011 Art. Nr. 90 32 025
9
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.6 Ändern der Basiseinstellungen
Im Menü Basiseinstellungen Erweiterte Funktionen ( b.EF
) werden die allgemeingültigen
Basiseinstellungen vorgenommen. Diese Einstellungen beeinflussen die Darstellung in der
Messwertanzeige ebenso wie die Einstellmöglichkeiten in den Menüs Füllstand und Temperatur. Auch die Zuordnung der Schaltausgänge zur Füllstands- bzw. Temperaturmessung (soweit vorhanden) kann hier geändert werden.
¾ Drücken Sie die Taste , um ins Hauptmenü zu gelangen.
¾ Wählen Sie den Menüpunkt b.EF
mit den Tasten und aus und öffnen Sie das Menu mit der
Taste .
4.6.1 Einheit Temperatur festlegen
Hier wird das angezeigte Einheitensymbol für die Temperatur festgelegt.
Folgende Wahlmöglichkeiten stehen zur Verfügung:
Grad Celsius Grad Fahrenheit
Hinweis:
-
Bei Änderung der Einstellung werden alle zugehörigen Einstellungen wie z.B. Einstellung der Schaltpunkte entsprechend geändert.
Bei der Installation mit Fernanzeige müssen zur genauen Messung der Temperatur sowohl die Länge der
Zuleitung als auch der Querschnitt eingegeben werden, um den Leitungswiderstand rechnerisch zu kompensieren.
Die Leitungslänge sollten in folgendem Menü angepasst werden:
Einstellbereich
0 m ... 20 m
Länge des Messkabels zur Kompensation des Zuleitungswiderstandes Pt100
Hinweis:
-
Da Hin- und Rückleitung berücksichtigt werden müssen, ist die doppelte
Kabellänge einzugeben!
Geben Sie hier den Leitungsquerschnitt der Zuleitung ein.
10
Einstellbereich
0.25|| bis 1.5||
Querschnitt des Messkabels zur Kompensation des Zuleitungswiderstandes
Pt100
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.6.4 Aktualisierungsrate der Anzeige einstellen
Je nach Anwendung kann die Aktualisierungsrate der Anzeige eingestellt werden. Die Anzeige kann auch ganz ausgeschaltet werden. Die Funktion der LEDs ist weiterhin gegeben.
Folgende Auswahlmöglichkeiten stehen zur Verfügung: schnell medium langsam Anzeige aus
Hinweis:
− Fehlermeldungen werden trotz ausgeschaltetem Display angezeigt.
4.6.5 Tastensperre aktivieren / deaktivieren
Um unbefugte Änderungen der Einstellungen im Gerät zu verhindern, kann eine Tastensperre eingerichtet werden.
Die Tastensperre wird aktiviert, wenn mindestens eine Ziffer > 0 eingegeben wird. Während der Eingabe wird die aktive Ziffer mit einem Punkt markiert.
Einstellbereich:
000 bis 999
¾ Öffnen Sie die Werteliste mit der Taste :
¾ Stellen Sie die Ziffer mit den Tasten und ein (0 bis 9) und bestätigen
Sie mit der Taste . Die aktive Ziffer rückt um eine Stelle nach rechts.
¾ Bestätigen Sie den Code zum Schluss mit der Taste .
Das Gerät kehrt zum Untermenü zurück.
Hinweis:
-
Tastatursperre aufheben mit der Eingabe: 000
4.6.6 Werkseinstellungen wieder herstellen (Reset)
Mit der Funktion Reset ( rES ) können die Werkseinstellungen wieder hergestellt werden. Dabei gehen alle
Änderungen verloren. Da die Grenzwerte ebenfalls zurückgesetzt werden, müssen die Einstellungen für den Füllstand und die Temperatur unbedingt überprüft werden.
Folgende Wahlmöglichkeiten stehen zur Verfügung:
Auslieferungszustand: Nein, die aktuellen Einstellungen werden beibehalten
Die Werkseinstellungen lauten folgendermaßen:
BX140013, 02/2011 Art. Nr. 90 32 025
Auslieferungszustand: Ja die Einstellungen werden auf die werkseitigen
Standard-Einstellungen zurückgesetzt.
11
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
Definitionen:
SPx / rPx Schaltpunkt / Rückschaltpunkt x dsx / drx Einschaltverzögerung / Rückschaltverzögerung für Schaltausgang x
Ax.Hi / Ax.Lo maximaler und minimaler Messwert für die Ausgabe
A.oux
Signalform des Analogausgangs oux Schaltcharakteristik des Schaltausgangs x t.uni
r.oux
Einheit für Temperatur
Zuordnung des Schaltausgangs x zur Temperaturüberwachung dis Aktualisierungsrate des Displays
Loc Tastensperre
SJ.ou
protokollierter dt.MM
Verzögerung zur Aufzeichnung der minimalen / maximalen Temperatur
Hinweis: Bei kundenspezifischen Vorgaben kann die werkseitige Voreinstellung von den hier aufgeführten
Werten abweichen.
Ausführung mit 2 Schaltausgängen:
Diagnose Schaltausgänge
SP1
/ rP1 ds1
/ dr1
/ ou1
SP2
/ rP2 ds2
/ dr2
/ ou2
0
0
50
/
/
0
60
0
/
/
/
/
45
Hno
55
Hno
Basiseinstellungen
t.uni r.ou1 r.ou2 dis
Loc
∞C tEMP tEMP
FASt
000
Ausführung mit 4 Schaltausgängen:
Schaltausgänge
SP1
/ rP1 ds1
/ dr1
/ ou1 0
50
/
0
/
/
45
Hno
SP2
/ rP2 ds2
/ dr2
/ ou2
SP3
/ rP3 ds3
/ dr3
/ ou3
SP4
/ rP4 ds4
/ dr4
/ ou4
0
0
60
/
0
70
/
/
/
55
Hno
65
0
/
0
/
Hno
80
/
75
/
0
/
Hno
Basiseinstellungen
t.uni r.ou1 r.ou2 r.ou3 r.ou4 dis
Loc
∞C tEMP tEMP tEMP tEMP
FASt
000
Ausführung mit 1 Schaltausgang und 1 Analogausgang:
SP1
/ rP1 ds1
/ dr1
/ ou1
A1.Hi
/
A1.Lo
/
Schaltausgang
Analogausgang
A.ou1
0
50
/
45
0
/
0
/
Hno
/
100
/ i1
Basiseinstellungen
t.uni r.ou1
∞C tEMP dis
Loc
FASt
000
SJ.ou dt.MM
SJ.ou dt.MM
SJ.ou dt.MM
Diagnose
out1
0.0 out1
0.0
Diagnose
out1
0.0
12
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.7 Schaltausgänge
Alle Schaltausgänge werden in gleicher Weise eingestellt. Die Nummer des Schaltausgangs wird daher mit x dargestellt. Rufen Sie den einzustellenden Schaltausgang über das Menü der entsprechenden
Messgröße auf.
Im Untermenü Erweiterte Funktionen können weitere Einstellungen für jeden einzelnen Schaltausgang vorgenommen werden, die z. B. das Schaltverhalten des Ausgangs beeinflussen. Auch ein Test des
Ausgangs ist hier möglich.
4.7.1 Schaltausgang x: Definition der Schaltcharakteristik
Die Schaltcharakteristik für den Ausgang wird in folgendem Menü festgelegt:
Folgende Wahlmöglichkeiten stehen zur Verfügung:
Hysteresefunktion
Fensterfunktion als
Schließer
Hysteresefunktion als Schließer
Fensterfunktion
Fensterfunktion als
Öffner
Hysteresefunktion als Öffner
Schließer- oder Öffner-Funktion, bei der beim Überschreiten des eingestellten Schaltpunktes das Ausgangssignal gesetzt wird. Beim Unterschreiten des eingestellten Rückschaltpunktes wird das Ausgangssignal gelöscht.
SP rP
Schließer
bedeutet hier, dass der
PNP- Schaltausgang oberhalb des Hno
(Schließer)
Schaltpunktes SPx geschlossen ist und bei Unterschreiten des Rückschaltpunktes rPx wieder öffnet.
Hnc
(Öffner)
0
Öffner
bedeutet hier, dass der PNP-
Schaltausgang oberhalb des Schaltpunktes SPx geöffnet ist und bei
Unterschreiten des Rückschaltpunktes rPx wieder schließt.
0
1
1
Siehe auch die Erläuterungen in der
Zeichnung unten.
Schließer- oder Öffner-Funktion, mit der ein Signalfenster festgelegt wird.
Bei Erreichen des Messfensters wird das Ausgangssignal gesetzt, beim
Verlassen wieder gelöscht.
Schließer
bedeutet hier, dass der
PNP- Schaltausgang geschlossen ist, wenn sich der Wert innerhalb des Fensters befindet. Sonst ist der
Schaltausgang geöffnet.
FH
FL
Fno
( Schließer)
1
0
Fnc
( Öffner)
1
0
Öffner
bedeutet hier, dass der PNP-
Schaltausgang geöffnet ist, wenn sich der Wert innerhalb des
Fensters befindet. Sonst ist der
Schaltausgang geschlossen.
BX140013, 02/2011 Art. Nr. 90 32 025
Hysterese
SP rP )
Zeit t
Zeit t
F000573
Zeit t
Fenster
Zeit t
Zeit t
F000574
Zeit t
13
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
Frequenzausgang
Ist der Ausgang als Frequenzausgang definiert, wird ein Rechtecksignal mit einer Frequenz zwischen
1 Hz und 100 Hz proportional zum
Messwert ausgegeben.
Frequenzausgang
Hinweis: Um die Flankensteilheit des Rechtecksignals zu erhöhen, wird empfohlen, den Schaltausgang mit einem 10 k
Ω Widerstand zu belasten.
+24 V
+24 V
+24 V
Beispiel: F1.Lo
= 15 °C, F1.Hi
= 80 °C mit Temperatur T und Frequenz f:
T F1.Lo
entspricht f = 1 Hz
Zeit t
T = 47,5 °C entspricht f = 50 Hz
Zeit t
F000575
T F1.Hi
entspricht f = 100 Hz
Zeit t
Hinweis: Die Schaltfunktion kann unterschiedlich bezeichnet sein:
SP
ϑ
SP rP rP offen
(0) geschlossen
(1) offen
(0) geschlossen
(1)
14 steigend Schließer fallend Öffner
NO (normally open)
Hno steigend Öffner fallend Schließer
NC (normally closed)
Hnc
4.7.2 Schaltausgang x: Obere Schaltgrenze (Einschaltpunkt)
Die obere Schaltgrenze für den Schaltausgang Out X wird in folgendem Untermenü gesetzt:
Messgröße
Hauptmenü Messgröße
►
Out X
Einstellmenü OUTX
►
Obere Schaltgrenze X
►
Wahlmenü
Schaltpunkt für OUT x
Einstellbereich
0 °C bis 100 °C
(32 °F bis 212 °F)
Hinweis:
-
Der Schaltpunkt muss innerhalb der Bereichsgrenzen (siehe Menü
Basiseinstellungen Erweiterte Funktionen
) gewählt werden.
-
Falls dem Schaltausgang OUT 1 die Funktion Fenster zugewiesen wurde, wird
Fenstergrenze.
angezeigt. Der eingestellte Wert entspricht der oberen
-
Falls dem Schaltausgang OUT 1 die Funktion Frequenzausgang zugewiesen wurde, wird entspricht der Frequenz 100 Hz.
angezeigt. Der eingestellte Wert
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.7.3 Schaltausgang x: Untere Schaltgrenze (Rückschaltpunkt)
Die untere Schaltgrenze für den Schaltausgang Out 1 wird in folgendem Untermenü gesetzt:
Messgröße
Hauptmenü Messgröße
►
Out X
Einstellmenü OUTX
►
Untere Schaltgrenze X
►
Wahlmenü
Rückschaltpunkt für OUT 1
Einstellbereich
0 °C bis 100 °C
(32 °F bis 212 °F)
Hinweis:
-
Der Rückschaltpunkt muss innerhalb der Bereichsgrenzen gewählt werden.
-
Falls dem Schaltausgang OUT 1 die Funktion Fenster zugewiesen wurde, wird
Fenstergrenze.
angezeigt. Der eingestellte Wert entspricht der oberen
-
Falls dem Schaltausgang OUT 1 die Funktion Frequenzausgang zugewiesen wurde, wird entspricht der Frequenz 1 Hz.
angezeigt. Der eingestellte Wert
4.7.4 Schaltausgang x: Einschaltverzögerung
Im Menü Erweiterte Funktionen EFx können weitere Einstellungen für den Schaltausgang x vorgenommen werden. Das Untermenü finden Sie auf der zweiten Untermenüebene:
Die Schalt- und Rückschaltverzögerungszeit verhindert das zu häufige Ansprechen des Alarms bei unruhigen Verhältnissen. Die Schaltverzögerung wird in folgendem Menü eingestellt:
Zeitspanne in Sekunden, in der das Signal kontinuierlich anliegen muss, damit der Schaltausgang anspricht.
Einstellbereich:
0…100 Sekunden
Hinweis:
-
Falls dem Schaltausgang OUT x die Funktion Fenster zugewiesen wurde, entspricht der eingestellte Wert der Einschaltverzögerung, die ein gültiges
Erreichen des Messfensters detektiert.
-
Falls dem Schaltausgang OUT x die Funktion Frequenzausgang zugewiesen wurde, hat dieser Wert keine Auswirkung.
4.7.5 Schaltausgang x: Rückschaltverzögerung
Die Rückschaltverzögerung wird in folgendem Menü eingestellt:
Verzögerung des Rückschaltsignals für OUT x
Zeitspanne in Sekunden, in der das Signal kontinuierlich anliegen muss, damit der Schaltausgang anspricht.
Einstellbereich:
0…100 Sekunden Hinweis:
-
Falls dem Schaltausgang OUT 1 die Funktion Fenster zugewiesen wurde, entspricht der eingestellte Wert der Einschaltverzögerung, die ein gültiges
Verlassen des Messfensters detektiert.
-
Falls dem Schaltausgang OUT 1 die Funktion Frequenzausgang zugewiesen wurde, hat dieser Wert keine Auswirkung.
BX140013, 02/2011 Art. Nr. 90 32 025
15
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.7.6 Schaltausgang x: Testen des Schaltausgangs
Ein Test des Schaltausgangs kann in folgendem Menü gestartet werden:
Messgröße
Hauptmenü Messgröße
►
Out X
Einstellmenü OUTX
►
EF X
Erweiterte Funktionen
►
Check Out X
Funktionstest OUTX
►
Wahlmenü
Testmöglichkeit für den Schaltausgang
Wahlmöglichkeiten bei Einstellung von ou1 auf Hno / Hnc / Fno / Fnc :
Normalbetrieb des
Schaltausgangs
Schaltausgang dauerhaft ausschalten
Wahlmöglichkeiten bei Einstellung von ou1 auf FM
Schaltausgang dauerhaft einschalten
Normalbetrieb als
Frequenzausgang
Ausgabe
Frequenz 1 Hz
Ausgabe
Frequenz 100 Hz
Hinweis:
-
Stellen Sie die Funktion nach Beendigung des Tests unbedingt auf
Normalbetrieb
n.oP
ein.
4.7.7 Anzeigefunktion der Status-LED ändern
Der Schaltzustand des Ausgangs wird durch die LEDs im Display signalisiert. Die Zuordnung der LED zum
Schaltausgang geht aus folgender Tabelle hervor:
Nummerierung LED Schaltausgang x Zuordnung bei
1 Schaltausgang
Zuordnung bei
2 Schaltausgängen
Zuordnung bei
4 Schaltausgängen
LED
1 2 3 4
3
4
1
2
LED 1 – gelb LED 1 – gelb
LED 2 - rot
LED 1 – gelb
LED 2 - rot
LED 3 – gelb
LED 4 - rot
In der Werkseinstellung zeigt die LED den physikalischen Zustand des PNP-Schaltausgangs an
(Schaltausgang geschlossen – LED leuchtet).
Möglicherweise soll die logische Funktion der Anzeige anders erfolgen als das physikalische Signal auf dem Schaltausgang. Sie können daher diese Anzeige in diesem Menüpunkt auch umkehren
(Schaltausgang geöffnet – LED leuchtet).
Beispiel anhand der Temperatur:
Sie haben 2 Schaltausgänge für die Temperatur, die folgendermaßen eingestellt sind:
• oberer Schaltkontakt: Max-Kontakt, steigend Schließer. Die LED leuchtet, wenn der Maximalwert der
Temperatur überschritten wird und die Temperatur außerhalb des gewünschten Bereiches ist. Der
Angezeigte Status ist also „Fehler“ bei Leuchten der LED.
• Unterer Schaltkontakt: Min-Kontakt, steigend Schließer. Die LED leuchtet also bei der
Werkseinstellung, wenn der Minimalwert der Temperatur überschritten wird. Es würde also in diesem Fall die LED leuchten, wenn der Status in Ordnung ist.
Die Tabelle zeigt ein Beispiel mit der Werkseinstellung und mit invertierter Statusfunktion für LED3. Die
Schaltpunkte sind wie folgt definiert:
16
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
SP3 = 70 °C, rP3 = 65 °C
SP4 = 80 °C, rP4 = 75 °C
Werkseinstellung Statusfunktion invertiert
Zustand Status
A
Temperatur steigt auf > 70 °C
PNP- Schaltausgang 3 geschlossen
OK
LED3 AN LED3 AUS
B
Temperatur steigt auf > 80 °C
PNP- Schaltausgang 4 geschlossen
Fehler
LED4 und LED3 AN nur LED4 AN
C
Temperatur fällt auf < 75 °C
PNP- Schaltausgang 4 geöffnet
OK
LED3 AN LED3 AUS
D
Temperatur fällt auf < 65 °C
PNP- Schaltausgang 3 geöffnet
Fehler
LED3 AUS LED3 AN
Hier können Sie für einen Kontakt die Statusfunktion der LED umkehren: die LED leuchtet, wenn der
Kontakt geöffnet ist, also unterhalb der minimalen Temperatur, und es wird bei Leuchten der LED wieder der Status „Fehler“ angezeigt. Insbesondere hängt das Aufzeichnen von Ereignissen vom Aufleuchten der
LED ab (siehe Kapitel „Diagnosemöglichkeiten“, 0).
Folgende Wahlmöglichkeiten stehen zur Verfügung
LED = output; die LED leuchtet, wenn der PNP-
Schaltausgang geschlossen ist.
LED = -output; die LED leuchtet, wenn der PNP-
Schaltausgang geöffnet ist.
Hinweis:
-
Insbesondere hängt das Aufzeichnen von Ereignissen vom Aufleuchten der
LED ab (siehe Kapitel „Diagnosemöglichkeiten“, 0).
BX140013, 02/2011 Art. Nr. 90 32 025
17
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.8 Analogausgänge
4.8.1 Analogausgang x: Zuweisung der oberen Grenze
Hier erfolgt die Zuweisung, bei welcher Temperatur das maximale Analogsignal ausgegeben werden soll.
Die Einstellung erfolgt im Menü
Messgröße
Hauptmenü Messgröße
►
Analog OutX
Einstellmenü Analog X
►
Analog MaxX
Zuweisung Maximalwert
►
Wahlmenü
Einstellbereich
0 °C bis 100 °C
(32 °F bis 212 °F)
Hinweis:
-
Der eingestellte Ausgabebereich darf nicht kleiner als 10% des
Messbereichs gewählt werden: A1.Hi
– A1.Lo
>= 10%
-
Bei zu klein gewähltem Bereich kann es sein, dass die
Analogwertausgabe Stufen aufweist.
4.8.2 Analogausgang x: Zuweisung der unteren Grenze
Hier erfolgt die Zuweisung, bei welcher Temperatur das minimale Analogsignal ausgegeben werden soll.
Die Einstellung erfolgt im Menü
Messgröße
Hauptmenü Messgröße
►
Analog OutX
Einstellmenü Analog X
►
Analog MinX
Zuweisung Minimalwert
►
Wahlmenü
Einstellbereich:
0 °C bis 100 °C
(32 °F bis 212 °F)
Hinweis:
-
Der eingestellte Ausgabebereich darf nicht kleiner sein als 10% des
Messbereichs gewählt werden: A1.Hi
– A1.Lo
>= 10%
-
Bei zu klein gewähltem Bereich kann es sein, dass die
Analogwertausgabe Stufen aufweist.
4.8.3 Analogausgang x: Festlegen der Signalform
Der Analogausgang kann als Spannungs- oder Stromausgang mit unterschiedlichen Wertebereichen definiert werden. Die Einstellung erfolgt im Menü
Folgende Wahlmöglichkeiten stehen zur Verfügung:
4 mA bis 20 mA 2 V bis 10 V 0 V bis 10 V 0 V bis 5 V
18
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.8.4 Analogausgang x: Testen des Analogausgangs
Auch der Analogausgang kann getestet werden. Der größte, der mittlere und der kleinste Analogwert können nacheinander ausgegeben werden. Die Einstellung erfolgt im Menü
Folgende Wahlmöglichkeiten stehen zur Verfügung:
Normalbetrieb Ausgabe höchster Analogwert
Ausgabe mittlerer Analogwert Ausgabe niedrigster Analogwert
Hinweis:
-
Stellen Sie die Funktion nach Beendigung des Tests unbedingt auf
Normalbetrieb
n.oP
ein.
4.9 Diagnosemöglichkeiten
Das Gerät ist in der Lage, die Ereignisse für einen Schaltausgang zu protokollieren. Als Ereignis wird dabei das Aufleuchten der LED bezeichnet. Damit hängt die Aufzeichnung der Schaltvorgänge von der
Einstellung der Schaltfunktion der LEDs ab (siehe 4.7.7).
Die Einstellungen und die Auswertung können hier vorgenommen werden.
!
HINWEIS
Es kann nur ein Schaltausgang protokolliert werden. Der zu protokollierende
Schaltausgang wird im Menüpunkt Schaltausgang Log Alarm S.LA
gesetzt.
¾
Drücken Sie die Taste , um ins Hauptmenü zu gelangen.
¾ Wählen Sie den Menüpunkt diA mit den Tasten und aus.
Von hier aus gelangt man zu diversen Diagnosewerten und Protokollierungen zur Füllstands- und Temperaturüberwachung.
¾ Öffnen Sie das Menü mit der Taste .
Nun können Sie die Diagnose-Einstellungen ändern bzw. aufrufen.
BX140013, 02/2011 Art. Nr. 90 32 025
19
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
Die letzten 6 Ereignisse des protokollierten Schaltausgangs können hier abgerufen oder gelöscht werden.
Die Journaleinträge werden in der folgenden Weise angezeigt:
Ereignis fand vor x Stunden (h) / Tagen (d) statt,
• Ereignisse 2 bis 5 fanden vor x Stunden / Tagen statt,
Ereignis fand vor x Stunden / Tagen statt
• Löschfunktion
--).
* noch nicht belegt, es fanden erst 4 Ereignisse statt
Beispiel:
Jor1
⇔
1.3h
,Taste
Jor2
⇔
2.4h
, Taste ,
Jor3
⇔
6.1h
, Taste ,
Jor4
⇔
8.2h
, Taste ,
Jor5
⇔ non *, Taste ,
Jor6
---
⇔ non *, Taste ,
, Taste ; = löschen
Der Index des Eintrags x wird im Wechsel mit der Zeit angezeigt, z.B.
Jor1
⇔
1.4h
für jüngstes Ereignis vor 1.4 Stunden.
Bestätigen der Anzeige und kehrt zum Untermenü zurück.
mit der Taste löscht die Ereignisliste
Hinweis:
− Falls keine Ereignisse aufgezeichnet wurden, wechselt die Anzeige zwischen Jorx und non .
− Die gespeicherten Daten werden nach 6 Monaten überschrieben.
4.9.2 Maximale und minimale Temperatur
Hier werden die gespeicherte maximale und minimale Temperatur angezeigt oder gelöscht.
Die Werte werden in der folgenden Weise angezeigt:
• Maximalwert der Temperatur,
• erreicht vor x Stunden / Tagen,
• Minimalwert der Temperatur,
• erreicht vor x Stunden / Tagen,
• Löschfunktion.
Beispiel:
72;C , Taste
8.4h
, Taste ,
22;C , Taste ,
2.1h
, Taste ,
--, Taste ; = löschen
20
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
Reihenfolge Menü:
Max. Wert,
Zeit
Min. Wert,
Zeit löschen (Reset)
Bestätigen der Anzeige und kehrt zum Untermenü zurück.
mit der Taste löscht die Ereignisliste
Hinweis:
− Die gespeicherten Daten werden nach 6 Monaten überschrieben.
4.9.3 Zu protokollierenden Schaltausgang festlegen
Hier wird der zu protokollierende Schaltausgang ausgewählt. Es kann nur ein Schaltausgang protokolliert werden.
Auswahl: out1 bis outX
Hinweis:
− Das Sichern der Werte vom flüchtigen in den nicht flüchtigen Speicher erfolgt ca. alle drei Stunden.
− Die gespeicherten Daten werden nach 6 Monaten überschrieben.
4.9.4 Verzögerung zur Min/Max-Speicherung der Temperatur
Um bei Temperaturschwankungen zuverlässige Werte aufzuzeichnen, kann eine Verzögerungszeit zur
Speicherung der minimalen und der maximalen Temperatur eingestellt werden. Hier wird die Zeitspanne in
Sekunden angegeben, in der das Signal kontinuierlich anliegen muss, ehe die Temperatur protokolliert wird.
Stellen Sie die gewünschte Verzögerung ein.
Einstellbereich:
0…100 Sekunden
BX140013, 02/2011 Art. Nr. 90 32 025
21
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
5 Instandhaltung
5.1 Inspektion und Wartung
Das Gerät arbeitet wartungsfrei.
5.2 Service und Reparatur
Sollte ein Fehler beim Betrieb auftreten, finden Sie in der folgenden Tabelle Hinweise für die Fehlersuche und Beseitigung.
Sollten Sie Fragen haben, wenden Sie sich bitte an unseren Service
Tel.: +49-(0)2102-498955 oder Ihre zuständige Vertretung. Halten Sie dazu bitte die Daten des
Typenschildes bereit.
Ist nach Beseitigung eventueller Störungen und nach Einschalten der Netzspannung die korrekte Funktion nicht gegeben, muss das Gerät durch den Hersteller überprüft werden. Bitte senden Sie das Gerät zu diesem Zweck in geeigneter Verpackung an:
Bühler Technologies GmbH
- Reparatur/Service -
Harkortstraße 29
40880 Ratingen
Deutschland
5.3 Fehlerbehebung
Im Fehlerfall werden alle Ausgänge spannungslos geschaltet. Die vier LEDs blinken. Fehler bleiben bis zum Ausschalten im Gerät gespeichert.
Problem / Störung mögliche Ursache Abhilfe
keine Anzeige − Kabel prüfen und ggf. austauschen
Fehlermeldungen im Display: Wechsel zwischen Err und Exxx: z.B.
⇔
− Umgebungstemperatur zu niedrig
Error 01
− Umgebungstemperatur zu hoch
Error 02
Error 04
Error 08
− Pt 100 defekt (Kurzschluss)
− Pt 100 defekt (Kabelbruch)
− Zuleitung Pt 100 austauschen
− Gerät zur Reparatur einsenden
− Zuleitung Pt 100 austauschen
− Gerät zur Reparatur einsenden
6 Entsorgung
Bei der Entsorgung sind die gesetzlichen Vorschriften des Anwenderlandes zu beachten, insbesondere die
Vorschriften für die Entsorgung von elektronischen Bauteilen.
7 Beigefügte Dokumente
-
Konformitätserklärung
22
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
1 Introduction
The Thermotronik TT77 is designed for temperature monitoring in fluid systems. It is equipped with up to four switching outputs or one switching output combined with an analog output. Tank-top installation of the display and control unit (together with the switching tube, TT77F…) is provided as well as 35 mm DIN rail mounting as remote display (TT77W…). If the display is mounted directly on the tank, it can be rotated by
270°.
The switching characteristic of the output is adjustable as window, hysteresis or frequency output. A menu based on VDMA standard 24574 provides easy programming. Minimum and maximum temperature as well as six events can be stored (logbook).
Please find the configuration on the type plate. Here you find the order number, the product key and the model description.
TT77 with integrated display
TT-77F-
A
L
T
E
R
N
Design
MS
VA
-2T
-4T
-1T-KT
-2T-KT
-4T-KT
brass stainless steel
2 x PNP output
4 x PNP output
1 x PNP output /
2 x PNP output /
4 x PNP output /
1 x Analogue output
1 x Analogue output
1 x Analogue output
Length
280
(max. 1000 mm)
370
500
Variable (please specify)
TT77 with remote display
TT-77W-
Design
MS
VA
brass stainless steel
Length
280
(max. 1000 mm)
370
500
Variable (please specify)
A
L
T
E
-2T
-4T
-1T-KT
-2T-KT
-4T-KT
2 x PNP output
4 x PNP output
1 x PNP output / 1 x Analogue output
2 x PNP output /
4 x PNP output /
1 x Analogue output
1 x Analogue output
WARNING
All types of devices are designed for industrial applications only. They are not designed as safety devices. They must not be used if in case of failure or malfunction safety or health of persons is affected.
The device is not suitable for use in hazardous areas with potentially explosive atmosphere.
BX140013, 02/2011 Art. Nr. 90 32 025
23
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
Please check prior to installation of the device that the technical data matches the application parameters.
Check that the delivery is complete as well.
Operation of the device is only valid if
-
the product is used under the conditions described in the installation- and operation instruction, the intended application according to the type plate and the intended use. In case of unauthorized modifications done by the user Bühler Technologies GmbH can not be held responsible for any damage .
-
the performance limits given in the datasheets and in the installation- and operation instruction are obeyed,
-
monitoring devices and safety devices are installed properly,
-
service and repair is carried out by Bühler Technologies GmbH, unless described in this manual,
-
only original spare parts are used.
This manual is part of the equipment. The manufacturer keeps the right to modify specifications without advanced notice. Keep this manual for later use.
The following warning signs and signal words are used in this manual:
Warning against hazardous situation
Warning against respiration of toxic gases
Warning against acid and corrosive substances disconnect from mains wear respirator wear face protection
Warning against possible explosive atmospheres
Signal words for warnings:
Signal word for important information to the product.
NOTE
CAUTION
wear gloves
Signal word for a hazardous situation with low risk, resulting in damaged to the device or the property or minor or medium injuries if not avoided.
WARNING
Signal word for a hazardous situation with medium risk, possibly resulting in severe injuries or death if not avoided.
DANGER
Signal word for an imminent danger with high risk, resulting in severe injuries or death if not avoided.
Installation of the device shall be performed by trained staff only, familiar with the safety requirements and risks.
Adhere to all relevant safety regulations and technical indications for the specific installation place. Prevent failures and protect persons against injuries and the device against damage.
The person responsible for the system must secure that:
-
safety and operation instructions are accessible and followed,
-
local accident prevention regulations and standards are obeyed,
-
performance data and installation specifications are regarded,
-
safety devices are installed and recommended maintenance is performed,
-
national regulations for disposal of electrical equipment are obeyed.
24
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
Maintenance and repair
-
Repairs on the device must be carried out by Bühler authorized persons only.
-
Only perform modifications, maintenance or mounting described in this manual.
-
Only use original spare parts.
During maintenance regard all safety regulations and internal operation instructions.
DANGER
Toxic and corrosive gases
Sample gas can be hazardous.
Take care that the gas is exhausted in a place where no persons are in danger.
Before maintenance turn off the gas supply and make sure that it cannot be turned on unintentionally.
Protect yourself during maintenance against toxic / corrosive gases. Use gloves, respirator and face protector under certain circumstances.
3 Installation and Connection
The device must be installed by trained personnel only. Regard all National and local safety regulations and internal operation instructions.
DANGER
Toxic and corrosive gases
Sample gas can be hazardous.
Take care that the gas is exhausted in a place where no persons are in danger.
Before maintenance turn off the gas supply and make sure that it cannot be turned on unintentionally.
Protect yourself during maintenance against toxic / corrosive gases. Use gloves, respirator and face protector under certain circumstances.
For direct tank top installation screw the switching tube into the mounting hole.
In case of installation with remote display, the display and control unit is mounted on a 35 mm DIN rail. The sensor is connected with a cable (max. length 10 m). The sensor is connected to the bottom of the display and control unit. Supply voltage and signal output are connected on the top.
The nominal supply voltage is 24 V DC. For pin assignment, please refer to the drawing in chapter 8.
The switching outputs are designed as PNP transistor outputs (see figure).
Note:
If the switching output is measured with high-impedance measuring equipment or if the frequency output is used, connect a 10 k
Ω resistor between output and ground to avoid faulty measurements.
24 V
+
-
F000588x
BX140013, 02/2011 Art. Nr. 90 32 025
25
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4 Operation
If an error message occurs during normal operation, please refer to chapter 5.2 “Troubleshooting”.
After connecting the device to the mains, the software version is displayed for a short time. Afterwards, the measured values are displayed.
LEDs
Display
Key ٛ
F000558X
Key ٛ
Key ٛ
Light emitting diodes above the reading indicate the switching state of the outputs. The LEDs are assigned to the switching output numbered the same. The following table shows the factory settings.
LED 1 – yellow indicates switching output 1
LED 2 - red indicates switching output 2 (if equipped)
LED 3 – yellow indicates switching output 3 (if equipped)
LED 4 – red indicates switching output 4 (if equipped)
The switching behavior of the LED (ON at closed or opened switching output) can be changed, see chapter
4.4.7.
26
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.3 Common Key Functions
The device is operated by three keys below the display.
Please refer to the chapters below for detailed descriptions.
Key Operation Mode:
- Normal mode:
Function
switches the displayed unit; example:
( )
- Within the menu: calls a sub-menu
- At the end of the menu: exits the sub-menu
+ *
(Exit) indicates the end of the menu
- After editing / selecting: accept and store the entered value or selection
- Normal mode:
- Within the menu: displays the configuration scrolls up menu item or selection, increase the displayed value. Holding the key changes the value continuously.
- Normal mode:
- Within the menu:
- Within the menu:
- Within the menu: calls the main menu scrolls down menu item or selection, decrease the displayed value. Holding the key changes the value continuously. exits the main or sub-menu and returns to normal mode exits the sub-menu
+ *
60 s no action * - Within the menu: exits the main or sub-menu
* Changed values are not stored if you exit the menu this way.
To select a menu item and to change a parameter value, follow the steps:
¾ Open the main menu with .
¾ Select the sub-menu using and and press to open the selected menu.
¾ If applicable, select the next sub-menu using and and press to open it.
¾ Select the menu item using and and press to show the parameter list or parameter value.
¾ Change the value using and and press to accept the changing. The changed parameter is stored and the device returns to the sub-menu.
¾ Exit the sub-menu by selecting menu item EXIT and press . The device returns to the subordinate menu or to the measurement display, respectively.
BX140013, 02/2011 Art. Nr. 90 32 025
27
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.4 Key Lock active
If the key lock is activated, pressing displays digit is indicated by a dot.
instead of opening the main menu. The active
¾ For each digit, enter the correct number using and and press . The active digit shifts to the right. After entering the third digit, the main menu opens.
If you entered a wrong password, the device returns to normal mode. If you have lost the password, enter the master code 287.
To unlock the key lock, select menu item Loc in sub-menu “Basic Settings Extended Functions” b.EF
and reset the password to 000.
4.5 Overview Menu Structure
The menu structure is based on VDMA standard 24574. The menu is structured hierarchically. The first level shows the main menu with menu item like tEMP , b.EF
, diA , E .
The menu items depend on the configuration. Your device may not provide all menu items described below. The configuration is displayed by pressing in normal operation mode. A 4-digit code is shown, for example: t s a v
F000583x
The four digits tsav have the following meaning: t: Type t = temperature measurement o = level and temperature measurement s: No. of switching outputs 1, 2 or 4 a: No. of analog outputs 0 or 1 v: Mounting / installation i = standard mounting (tank top installation)
F
►
Configuration
Device configuration
▲
Display
Measured values
▲▼
Temperature
Main menu temperature
▲▼
►
Basic EF
Basic ext. functions
Diagnostic
Main menu diagnostics
Exit
Exit submenu
►
Out 1
Settings for OUT1
▲▼
Out x
Settings for OUTx
▲▼
Analog Out1
Settings for Analog 1
▲▼
Exit
Exit submenu
If equipped
If equipped
The menu item is only displayed if the option is equipped. Example: if a = 0 than menu items for setting analog outputs are not available. You may skip these descriptions.
The structure of the menus for the switching outputs and analog outputs is similar. These menus include all settings for the switching outputs and the analog outputs
(if equipped).
General settings can be changed as well. They are set in main menu Basic settings extended functions ( b.ef
).
These settings should be changed first because they affect displays and settings of the sub-menus. General settings are for example used units and assignment of the switching outputs to level or temperature control. The assignment of the analog outputs cannot be changed.
In addition, menu Diagnostic provides methods for analyzing events.
Please refer to the detailed presentation of the menu structure at the end of this document.
28
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.6 Changing general Settings
In menu Basic Settings Extended Functions ( b.EF
) general settings are made. These settings affect the display of values as well as the options in menus Level and Temperature. This menu provides assignment of switching outputs (if equipped) as level or temperature measurement as well.
¾ Press to open the main menu.
¾ Select menu item b.EF
using and .
4.6.1 Setting Unit for Temperature Measurement
In this menu, the unit for temperature is set.
The following units are available:
Parameters:
[°C, °F] degree Celsius degree Fahrenheit
Note:
-
If the unit is changed, all related values as settings of the switching points are re-calculated.
4.6.2 Setting Cable Length
If the device is installed with remote display, length of the cable as well as cable cross section must be set to compensate the line resistance for exact temperature measurement.
The cable length should be adjusted in menu
Range:
0 m … 20 m length of the RTD measuring cable for compensating the line resistance
Note:
-
Enter twice the cable length in order to include supply and return line.
4.6.3 Setting Cable Cross Section
The cable cross section is set in menu
Range:
0,25 mm² … 1,5 mm² cross section of the RTD measuring cable for compensating the line resistance
BX140013, 02/2011 Art. Nr. 90 32 025
29
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.6.4 Setting the Display Refresh Rate
Depending on the application, the display refresh rate can be set. The display can be shut off as well. In this case, the function of the LED remains active.
The following options are available: fast medium slow
Note:
− Error messages are displayed even if the display is shut off. display off
4.6.5 Activating / deactivating Key Lock
To prevent unauthorized parameter changing, activate the key lock.
The key lock is active if at least one digit is set >0. When setting the password, the active digit is indicated with a dot.
Range:
000 to 999
¾ Open the parameter menu using .
¾ Enter first number of the password (0 to 9) using and and accept it by pressing . The active digit moves one step to the right.
¾ Enter all digits and accept the password with .
The device returns to the sub-menu.
Note:
-
Enter 000 to deactivate the key lock.
4.6.6 Reset Factory Settings
Reset
( rES ) resets the factory settings. All custom-made changes are lost. Since the limits are reset as well, all settings for level and temperature must be checked.
The following options are available:
No reset: keeps the actual settings.
Do reset: the settings are reset to factory settings.
30
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
The factory settings are set as follows:
Definitions:
SPx / rPx dsx / drx
Ax.Hi
/ Ax.Lo maximum and minimum measured value for output
A.oux
oux t.uni
Set point / reset point x
Delay for setting / resetting switching output x
Signal characteristics of the analog output signal switching characteristics for switching output x unit r.oux dis assignment of switching output x to level or temperature monitoring refresh rate of the display
Loc key
SJ.ou
dt.MM
recorded switching output delay for recording minimum / maximum temperature
Note: In case customized default values are set these values may differ from the standard factory settings as given in the following tables.
Design with 2 switching outputs:
Switching Outputs
SP1
/ rP1 ds1
/ dr1
/ ou1
SP2
/ rP2 ds2
/ dr2
/ ou2
0
0
50
/
/
0
60
0
/
/
/
/
45
Hno
55
Hno t.uni r.ou1 r.ou2 dis
Loc
Basic Settings
∞C tEMP tEMP
FASt
000
SJ.ou dt.MM
Diagnostics
out1
0.0
Design with 4 switching outputs:
A1.Hi
/
Switching Outputs
SP1
/ rP1 ds1
/ dr1
/ ou1
SP2
/ rP2 ds2
/ dr2
/ ou2
SP3
/ rP3 ds3
/ dr3
/ ou3
SP4
/ rP4 ds4
/ dr4
/ ou4
Switching Output
SP1
/ rP1 ds1
/ dr1
/ ou1
A1.Lo
/
A.ou1
0
0
0
0
50
/
0
60
/
70
/
80
/
0
0
0
/
Analog Output
/
/
/
/
/
/
/
0
45
Hno
55
Hno
65
Hno
75
Hno
0
/
0
/
Hno
/
50
/
45
100
/ i1 t.uni r.ou1
Basic Settings
∞C tEMP r.ou2 r.ou3 r.ou4 dis
Loc tEMP tEMP tEMP
FASt
000
Design with 1 switching output and 1 analog output: t.uni r.ou1
Basic Settings
∞C tEMP dis
Loc
FASt
000
SJ.ou dt.MM
Diagnostics
SJ.ou dt.MM out1
0.0
Diagnostics
out1
0.0
BX140013, 02/2011 Art. Nr. 90 32 025
31
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
All switching outputs are set in a similar way. Therefore, the number of the switching output is marked “X”.
Select the desired switching output by opening the respective main menu.
Sub menu Extended functions ( EF X) provides further settings for each switching output affecting for example the switching characteristics of the output. The switching output can be tested as well in this sub menu.
4.7.1 Switching Output 1: Definition of the Switching Characteristics
The switching characteristic is set in the following menu:
Meas. value
Main menu meas. value
►
Out X
Settings OUTX
►
EF X
Extenden functions
►
Set Out X
Definition OUTX
►
Parameter
The following options are available:
Hysteresis
Hysteresis,
NO contact
Hysteresis,
NC contact
Window
Window with
NO contact
Window with
NC contact
The switch works as NO or NC contact. If the set point is exceeded, the output signal is set. If the measuring signal falls below the reset point the output signal is reset.
SP rP
NO contact
is to be understood in the following way: the PNP switching output is closed if the measured value
Hno
( NO contact) is above switching point
SP opens if it falls below reset point x and rP x.
1
0
NO contact
is to be understood in the following way: the PNP switching output is open if the measured value is
Hnc
( NC contact)
1
0 above switching point
SP if it falls below reset point x and closes rP x. see as well figure below
The switch works as NO or NC contact. The limits of the window are set by FH and FL . If the measuring signal reaches the window the output signal is set. If the measuring signal leaves the window the output signal is reset.
NO contact
is to be understood in the following way: the PNP switching output is closed if the measured value is within the window limits. Otherwise it is open.
FH
FL
Fno
( NO contact)
1
0
(
Fnc
NC contact)
1
0
NC contact
is to be understood in the following way: the PNP switching output is open if the measured value is within the window limits. Otherwise it is closed.
Hysteresis
( SP rP )
Time t
Time t
F000573
Time t
Window
Time t
Time t
F000574
Time t
32
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
Frequency output
In this case, a square wave signal with a frequency between 1 Hz and
100 Hz in proportion to the measuring signal is set at the output.
Frequency output
Note: In order to increase the slope of the square wave, we recommend loading the switching output with a
10 k
Ω resistor.
+24 V
+24 V
+24 V
Example: F1.Lo
= 15 °C, F1.Hi
= 80 °C with temperature T and frequency f:
T F1.Lo
is equivalent to f = 1 Hz
Time t
T = 47,5 °C is equivalent to f = 50 Hz
Time t
F000575
T F1.Hi
is equivalent to f = 100 Hz
Time t
Note: The switching function may have different notations:
SP
ϑ
SP rP rP open
(0) closed
(1) open
(0) closed
(1) rising closer falling opener
NO (normally open)
Hno rising opener falling closer
NC (normally closed)
Hnc
4.7.2 Switching Output x: Upper Limit (Set Point)
The upper limit for switching output 1 (set point) is set in sub-menu:
Meas. value
Main menu meas. value
►
Out X
Settings OUTX
►
Set point X
►
Parameter
Range
0 °C to 100 °C
(32 °F to 212 °F)
Set point for OUT 1
Note:
-
The set point must be set within the measuring range (see menu Basic
extended functions
b.EF
).
-
If the switching characteristic is set to window,
The set value represents the upper window limit.
is displayed.
-
If the switching characteristic is set as frequency output, displayed. The set value represents a frequency of 100 Hz.
BX140013, 02/2011 Art. Nr. 90 32 025
is
33
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.7.3 Switching output x: Lower Limit (Reset Point)
The lower limit for switching output 1 (set point) is set in sub-menu:
Meas. value
Main menu meas. value
►
Out X
Settings OUTX
►
Restet point X
►
Parameter
Reset point for OUT 1
Range
0 °C to 100 °C
(32 °F to 212 °F)
Note:
-
The reset point rP1 must be set within the measuring range.
-
If the switching characteristic is set to window,
The set value represents the lower window limit.
is displayed.
-
If the switching characteristic is set as frequency output, displayed. The set value represents a frequency of 1 Hz. is
4.7.4 Switching Output x: Delay for Set Point
Menu Extended Functions
EF1
provides further settings for switching output 1. The menu is located on the second sub-menu level.
The delay time for set point and reset point avoids too many false alarms in case of fluctuating conditions.
The delay for the set point is set in the following menu:
Meas. value
Main menu meas. value
►
Out X
Settings OUTX
►
EF X
Extenden functions
►
Delay SP X
Delay for OUTX on
►
Parameter
Range:
0…100 seconds
Time in seconds which the signal must at least remain to enable the switch to operate.
Note:
-
If the output characteristic is set to Window the value represents the time delay for the signal remaining within the measurement window.
-
If the output characteristic is set to Frequency Output, this value has no effect.
4.7.5 Switching Output x: Delay for Reset Point
The delay for the set point is set in the following menu:
Meas. value
Main menu meas. value
►
Out X
Settings OUTX
►
EF X
Extenden functions
►
Delay RP X
Delay for OUTX off
►
Parameter
Range:
0…100 seconds
Delay for reset point OUT x
Time in seconds which the signal must at least remain to enable the switch to operate.
Note:
-
If the output characteristic is set to Window the value represents the time delay for the signal leaving the measurement window.
-
If the output characteristic is set to Frequency Output, this value has no effect.
34
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.7.6 Switching Output x: Testing the switching output
The switching output can be tested in the following menu:
Meas. value
Main menu meas. value
►
Out X
Settings OUTX
►
EF X
Extenden functions
►
Check Out X
Test for OUTX
►
Parameter
Testing the switching output
The following options are available if ou1 is set to Hno / Hnc / Fno / Fnc : normal operation turns the switching output off
The following options are available if ou1 is set to FM turns the switching output on normal operation as frequency output output frequency 1 Hz output frequency 100 Hz
Note:
-
After finishing the tests, make sure to reset the function to “normal
operation”
n.oP
.
4.7.7 Switching Output x: Changing Switching Function of LED
The indication of the switching output by the respective LED can be changed. The assignment of each LED is listed in the table.
Number of LED Switching output x 1 switching output 2 switching outputs 4 switching outputs
LED
1 2 3 4
1
2
3
4
LED 1 – yellow LED 1 – yellow
LED 2 - red
LED 1 – yellow
LED 2 - red
LED 3 – yellow
LED 4 - red
With factory settings, the LED indicates the physical switching state of the PNP output (output is closed –
LED is active).
If requested, it is possible to alter the logic function of the LED with respect to the physical signal of the output. Therefore, the function of the LED can be inverted in this menu item (output is opened – LED is active).
Example for temperature control:
Two switching outputs are available for temperature control with the following settings:
• Upper switching point: maximum, NO contact. The LED lights up if the maximum temperature is exceeded and temperature if off limits. The status indicated is “Error” if the LED in active (ON).
• Lower switching point: minimum, NO contact. The LED lights up if the minimum temperature is exceeded (factory setting). In this case, the LED would be active (ON) if no error is present (status
OK).
The following table shows an example with factory settings and inverted function for LED3. The switching points are defined as follows:
BX140013, 02/2011 Art. Nr. 90 32 025
35
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
SP3 = 70 °C, rP3 = 65 °C
SP4 = 80 °C, rP4 = 75 °C
Factory settings Inverted function of LED3
Situation Status
A
Temperature rises above 70 °C
PNP- switching output 3 is closed
OK
LED3 ON LED3 OFF
B
Temperature rises above 80 °C
PNP- switching output 4 is closed
Error
LED4 and LED3 ON only LED4 ON
C
Temperature falls below 75 °C
PNP- switching output 4 is closed
OK
LED3 ON LED3 OFF
D
Temperature falls below 65 °C
PNP- switching output 3 is opened
Error
LED3 OFF LED3 ON
You can invert the function of the LED for the respective switching output: The LED lights up if the contact is opened, or in other words if the temperature falls below the minimum temperature, and the active LED indicates an error. Especially the recording of events depends on the status of the LED (see chapter
“Diagnostic Tools” 4.9.1).
The following options are available:
LED = output; the LED lights up if the PNP switching output is closed.
LED = -output; the LED lights up if the PNP switching output is opened.
Note:
-
Recording of events depends on an activated LED lighting up (see chapter
Diagnostics, 4.9.1).
36
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.8.1 Analog Output x: Setting the upper Limit
Here, the upper limit of the measured temperature is assigned to the maximum analog output signal. The parameter is set in the following menu:
Range:
0 °C to 100 °C
(32 °F to 212 °F)
Note:
-
The set output range must not be smaller as 10% of the measuring range: A1.Hi
– A1.Lo
>= 10%
-
If the output range is set to small, the analog output signal may show steps.
4.8.2 Analog Output x: Setting the lower Limit
Here, the lower limit of the measured temperature is assigned to the minimum analog output signal. The parameter is set in the following menu:
Range:
0 °C to 100 °C
(32 °F to 212 °F)
Note:
-
The set output range must not be smaller as 10% of the measuring range: A1.Hi
– A1.Lo
>= 10%
-
If the output range is set to small, the analog output signal may show steps.
4.8.3 Analog Output x: Setting the Output Characteristic
The output characteristic of the analog output may be defined as voltage output or as current output with different signal output ranges. The output characteristic is set in the following menu:
The following options are available:
4 mA to 20 mA 2 V to 10 V 0 V to 10 V 0 V to 5 V
BX140013, 02/2011 Art. Nr. 90 32 025
37
38
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.8.4 Analog Output x: Testing the Analog Output
The analog output can be tested as well. The highest, the mean, and the lowest analog value can be put out one after the other. The parameters are set in the following menu:
The following options are available: normal operation output of max. analog value output of middle analog value
(e.g. 12 mA with 4-20 mA setting) output of min. analog value
Note:
-
After finishing the tests, make sure to reset the function to “normal
operation”
n.oP
.
The device is able to journalize the events of a selected switching output. An event is defined as an LED lighting up. Therefore, the recording of events depends on the settings for the switching function of the LED
(see chapter 4.7.7).
Settings and analysis are available in menu
!
NOTE
Only one switching output can be journalized. The respective switching output is selected in menu item “switching output journalize alarm log” S.LA
.
¾ Press to open the main menu.
¾ Select menu item diA using and .
This menu provides several diagnostic tools and journals of level and temperature measurement.
¾ Open the menu using .
Now you can set diagnostic parameters or recall the journal.
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.9.1 Open the Journal
The last 6 events of the switching output are listed and can be deleted.
The journal entries are listed as follows:
• Most recent event
Jor1 occurred x hours (h) / days (d) ago,
• Events 2 to 5 occurred x hours (h) / days (d) ago, event occurred x hours (h) / days (d) ago
• Delete
--).
Example:
Jor1
⇔
1.3h
, press
Jor2
⇔
2.4h
, press ,
Jor3
⇔
6.1h
, press ,
Jor4
⇔
8.2h
, press ,
Jor5
⇔ non *, press ,
Jor6
⇔ non *, press ,
--, press ; = delete
* not yet recorded; only 4 events occurred until now.
The display toggles between index of event x and the time it occurred, e.g.
Jor1
⇔
1.4h
recorded the latest event 1.4 hours ago.
Accepting with deletes the journal and returns to the sub-menu.
Note:
− If no event was recorded, the display toggles between
Jorx and non .
− The stored data is overwritten after six months.
4.9.2 Maximum and minimum Temperature
The recorded maximum and minimum level can be displayed or deleted.
The values are listed as follows:
• Maximum temperature,
• Reached x hours / days ago,
• Minimum temperature,
• Reached x hours / days ago,
• Delete
)
Accepting with deletes the journal and returns to the sub-menu.
Note:
− The stored data is overwritten after six months.
Menu order: max. value time min. value time delete (reset)
Example:
72;C , press
8.4h
, press ,
22;C , press ,
2.1h
, press ,
- - , press ; = delete
BX140013, 02/2011 Art. Nr. 90 32 025
39
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
4.9.3 Assigning the Switching Output for recording
In menu SJ.ou
the switching output to be recorded is assigned. Only one switching output can be journalized.
Range: out1 to outX
Note:
− Approx. every three hours the values of the volatile memory are transferred to the non-volatile memory.
− The stored data is overwritten after six months.
4.9.4 Delay for recording min/max Temperature
A delay time guarantees recording of reliable values for minimum and maximum levels in case of fluctuating temperature. A time interval (in seconds) is set, during which the signal must at least remain at maximum or minimum to be recorded.
Range:
0…100 seconds
Enter the desired delay time.
40
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
5 Maintenance
5.1 Inspection and Service
The device works maintenance free.
5.2 Service and Repair
If an error occurs during operation, the table below gives hints for troubleshooting.
If you need help or more information
call +49(0)2102-498955 or your local agent.
Please get the data of the type plate ready for identification.
If the device doesn’t work correctly after elimination of failures and turning power on, the device must be checked by the manufacturer. Please ship the device with suitable packing to
Bühler Technologies GmbH
- Reparatur/Service -
Harkortstrasse 29
40880 Ratingen
Germany
5.3 Troubleshooting
If an error occurs, all outputs are set idle. All four LEDs flash. Errors are stored until the device is shut off.
Problem / Failure Possible cause Solution
No display − Check cables and replace them if necessary
Error messages on the display: toggles between Err and Exxx: e.g.
Error 01
Error 02
Error 04
− Ambient temperature too low
⇔
− Keep limits according to technical data
− Ambient temperature too high − Keep limits according to technical data
− Pt 100 (RTD) defective (short cut) − Replace measurement cable
Pt 100 (RTD)
− Send the device for repair
Error 08
− Pt 100 (RTD) defective (broken wire)
− Replace measurement cable
Pt 100 (RTD)
− Send the device for repair
6 Disposal
Regard the local regulations for disposal of electric and electronic equipment.
-
Declaration of Conformity
BX140013, 02/2011 Art. Nr. 90 32 025
41
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
8 Technische Daten
Sonde Basiseinheit
max. Betriebsdruck
Betriebstemperatur
Material
Display
Temperaturanzeigebereich
Einstellbereich Alarm
8 Technical Data
Probe
5 bar / 14.5 psi max. operating press.
-40 °C … +100 °C / -40 °F … 212 °F operating temperature
Messing oder Edelstahl 1.4571 / brass or stainless steel 1.4571 material
-20 °C … 120 °C / -4 °F … 248 °F
0 °C … 100 °C / 32 °F … 212 °F
Display
temperature range alarm range
Auflösung
Schutzklasse Gehäuse
Einschaltstromaufnahme
Stromaufnahme im Betrieb
Versorgungsspannung
Nennspannung
Umgebungstemperatur
Ausgänge Temperatur
Temperatursensor
Stecker Sensoreingang Buchse
Stecker Ausgang Sockel
Programmierbare PNP
Transistorausgänge
Max. Schaltstrom pro Ausgang
Max. Schaltstrom Insgesamt
Analogausgang, programmierbar
0,5 °C / 1 °F
IP65
∼100 mA für / for 100 ms
∼ 50 mA
10 V … 32 V DC
24 V DC
-20 °C … 70 °C / -4 °F … 158 °F
Pt 100 Klasse B / RTD Class B
DIN EN 60751
M12, 4-pol.
M12, 4-pol. / M12, 8-polig
1 / 2 / 4
0.5 A
1 A
1
4-20 mA, 2-10 V, 0-10 V, 0-5 V resolution protection class housing inrush current operating current operating voltage nominal voltage ambient temperature
temperature outputs
temperature sensor connector sensor input , female
Connector output, male
Programmable PNP transistor outputs
Max. switching current per output
Max. switching current Over all analogue output, programmable
42
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
8.1 Abmessungen
Alle Abmessungen in mm / all dimensions given in mm
Gehäuse um min. 270° drehbar
Housing about min. 270° turnable
M12x1
8.1 Dimensions
M12x1
LED’s
LED’s
®
G1/2
®
SW 27 wrench size 27 mm
Eolastic gasket
Eolastic-
Dichtung bushing
M12x1
53
M12x1
SW 27 /
Rückseite / back view wrench size 27
Eolastic gasket
Eolastic -
Dichtung
G 1/2
40
53
BX140013, 02/2011 Art. Nr. 90 32 025
43
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
8.2 Steckerbelegung 8.2 Pin Assignment
Direktmontage / directly mounted
TT-77F…-2T TT-77F…-4T TT-77F…-1T-KT
3
4
5
2
6
8
7
1
3
2
4
1
3
2
4
1
TT-77F…-2T-KT TT-77F…-4T-KT
3
2
5
1
4
3
4
5
2
6
8
1
7
44
BX140013, 02/2011 Art. Nr. 90 32 025
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
Fernanzeige / remote display
TT-77W…-2T TT-77W…-4T
Temperaturfühler / temperature sensor Pt100
1
Pt 100
2 3 4
3
2
4
1
1
Pt 100
2 3 4
3
2
4
1
Sensoreingang / sensor input
2 3
2
4
1
2 3
2
4
1
Ausgang / output
Thermotronik
3
2
4
1
3
4
5
2
6
8
7
1
1
Pt 100
TT-77W…-1T-KT
2 3 4
3
2
4
1
3
2
4
1
2 3
2
4
1
Temperaturfühler / temperature sensor Pt100
1
Pt 100
TT-77W…-2T-KT TT-77W…-4T-KT
2 3 4
3
2
4
1
1
Pt 100
2 3 4
3
2
4
1
Sensoreingang / sensor input
2 3
2
4
1
2 3
2
4
1
Ausgang / output
Thermotronik
3
2
5
1
4
4
3
5
2
6
8
7
1
BX140013, 02/2011 Art. Nr. 90 32 025
45
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Thermotronik TT77
8.3 Aktuelle Einstellungen 8.3 Actual settings
Ausführung mit bis zu 4 Schaltausgängen / Design with up to 4 switching outputs:
Schaltausgänge
switching outputs
Basiseinstellungen
Basic settings
Diagnose
Diagnostic
SP1
/ rP1
* o.uni SJ.ou ds1
/ dr1
/ ou1 t.uni t.MM
SP2
/ rP2 ds2
/ dr2
/ ou2
SP3
/ rP3 o.Hi o.Lo dis dt.MM ds3
/ dr3
/ ou3
SP4
/ rP4
Loc ds4
/ dr4
/ ou4
Ausführung mit bis zu Schaltausgängen und bis zu 2 Analogausgängen /
Design with up to 2 switching outputs and up to 2 analog outputs:
SP1
/ rP1
Schaltausgänge
switching outputs
Basiseinstellungen
Basic settings
o.uni SJ.ou
Diagnose
Diagnostic
ds1
/ dr1
/ ou1 t.uni t.MM
SP2
/ rP2 o.Hi dt.MM o.Lo ds2
/ dr2
/ ou2
Analogausgänge
Analog outputs
A1.Hi
/
A1.Lo
/
A.ou1
dis
Loc
A2.Hi
/
A2.Lo
/
A.ou2
Datum:
Date:
Unterschrift:
Signature:
46
BX140013, 02/2011 Art. Nr. 90 32 025
EG-Konformitätserklärung
EC-declaration of conformity
Hiermit erklären wir, dass die nachfolgenden Produkte den wesentlichen Anforderungen der folgenden EG-
Richtlinie in ihrer aktuellen Fassung entsprechen:
Herewith we declare that the following products correspond to the essential requirements of the following EC
directive in its actual version:
2004/108/EG (EMV / EMC)
Produkte / products:
Temperaturmessgeräte und –anzeigen
Temperature sensors and display
Typ(en) / type(s):
Baureihe Thermotronik 77
Zur Beurteilung der Konformität wurden folgende harmonisierte Normen in aktueller Fassung herangezogen:
The following harmonized standards in actual revision have been used:
• EN 61326-1
Elektrische Mess-, Steuer-, Regel- und Laborgeräte - EMV-Anforderungen -
Teil 1: Allgemeine Anforderungen
Dokumentationsverantwortlicher für diese Konformitätserklärung ist der Unterzeichnende mit Anschrift am
Firmensitz.
The person authorised to compile the technical file is the one that has undersigned and is located at the company’s address
Ratingen, den 05. März 2010
__________________________________
Stefan Eschweiler
Technischer Leiter – technical manager
KX 11 0020
Bühler Technologies GmbH, Harkortstr. 29, D-40880 Ratingen,
Tel. +49 (0) 21 02 / 49 89-0, Fax. +49 (0) 21 02 / 49 89-20
Internet: www.buehler-technologies.com
9 Anhang: Übersicht Menüabfolge
48
49
9 Appendix: Overview Menu Structure
Main menu
Configuration
Device configuration
▲ ▼
Display
Measured values
▲ ▼
Temperature
Main menu temperature
►
Submenu 1
Pressing keys ▼+► or no action for 60 s:
Exits the menu
Out 1
Settings for OUT1
►
Submenu 2
Upper limit 1
▲ ▼
Lower limit 1
▲ ▼
EF 1
Extended functions
►
Submenu 3
►
Delay SP 1
Delay for set point
►
▲▼
▲▼
►
Delay RP 1
Delay for reset point
Set Out 1
Definition OUT1
Check Out1
Test of OUT1
LED Out1
Switching function LED1
Exit
Exit submenu
▲▼
►
Exit
Exit submenu
Out x
Settings for OUTx
Analog Out1
Settings for Analog 1
►
►
See menu structure Out1
Not available for models with analog output
►
▲▼
►
Analog Max1
Setting max. value
▲ ▼
Analog Min1
Setting min value
▲ ▼
Set A.OUt1
Setting characteristics
▲ ▼
Check An1
Test of Analog 1
▲ ▼
Exit
Exit submenu
►
►
►
►
Exit
Exit submenu
Parameters
Parameters
Parameters
►
Parameters
►
Parameters
►
Parameters
►
Parameters
►
Parameters
Setting for switching point / Setting of upper window limit
Defines value for frequency output 100 Hz (Out1 and Out2 only)
Setting for reset point / Setting of lower window limit
Defines value for frequency output 1 Hz (Out1 and Out2 only)
Time in seconds which the signal must at least remain to enable the switch to operate.
Setting range: 0...100 seconds (no function with frequency output)
Time in seconds which the signal must at least remain to enable the switch to reset.
Setting range: 0...100 seconds (no function with frequency output)
Defines the switching characteristics for output 1.
Hysteresis, NC or NO
Window, NC or NO
Frequency output 1-100 Hz
normal operation of the switching output /
Switching output OFF /
Switching output ON /
output 1 Hz
output 100 Hz
LED = output: same as switching output
LED = -output: inverted to switching output
Parameters
Parameters
Setting the maximum output signal to a defined value.
(Note: Max – Min >= 10% of the measuring range)
Setting the minimum output signal to a defined value.
(Note: Max – Min >= 10% of the measuring range)
Parameters
Parameters
Selecting the analog output signal:
= 4...20 mA normal operation
= 2...10 V output middle analog value
= 0...10 V = 0...5 V output maximum analog value output minimum analog value
Alternative menu for models with analog output
Not available for models with 2 or more switching outputs
50
Basic EF
Basis Erw. Funktionen
▲▼
Diagnostic
Diagnose Menü
▲▼
►
►
Reset
Restore factory settings
▲ ▼
Set Unit Temp
Unit of temperature
▲ ▼
Cable Length
Cable length PT100
▲ ▼
Cable Cross
Cross section PT100
▲ ▼
Display Rate
Display refresh rate
▲ ▼
Lock Device
Locking the keypad
▲ ▼
Exit
Exit submenu
►
Journal Out
List of alarms
▲ ▼
Max/Min Temp
Max-/Min memory temp.
▲ ▼
Set Jornal Out
Recorded output
▲ ▼
Delay MM
Delaty time for Min/Max
▲ ▼
Exit
Exit submenu
Exit
Exit menu
►
►
►
►
Alternative menu for models with remote display
Not available for models with integreted sensor
►
►
►
►
►
►
Confirm
Parameters
Parameters
Parameters
Parameters
Parameters
Resetting the device to factory settings
Selecting the displayed temperature unit.
[Celsius]
[Fahrenheit]
PT100 cable legth in meter (cable length X no. of strands), for compensating the cable resistance
ΔT[Δ°C]=0,462[Δ°C] x length[m]/cross section[mm²]
PT100 cable cross section in mm², for compensating the cabel resistance
Setting the display refresh rate
(schnell), (medium),
(langsam), (Anzeige aus)
Activating the key lock by setting a code >
Master code
Parameters
Parameters
Parameters
Parameters
Display of the last 6 alarms and reset.
Display: X – elapsed time in hours/days; scroll the journal with ▲▼
Display of min. and max. temperature and reset.
Display: value – elapsed time in hours/days; scroll with ▲▼
Selecting the output to be recorded
Time in seconds which the signal must at least remain to store minimum and maximum value.
Range: 0...100 seconds (no function with frequency output)
51
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 1 Bedienungs- und Installationsanleitung
- 4 1 Einleitung
- 4 1.1 Einsatzbereich
- 5 2 Wichtige Hinweise
- 6 3 Aufbauen und Anschließen
- 7 4 Betrieb
- 7 4.1 Einschalten
- 7 4.2 LED-Statusanzeigen
- 8 4.3 Allgemeine Tastenfunktionen
- 9 4.4 Aktive Tastensperre
- 9 4.5 Menü-Übersicht
- 10 4.6 Ändern der Basiseinstellungen
- 10 4.6.1 Einheit Temperatur festlegen
- 10 4.6.2 Zuleitungslänge festlegen
- 10 4.6.3 Zuleitungsquerschnitt festlegen
- 11 4.6.4 Aktualisierungsrate der Anzeige einstellen
- 11 4.6.5 Tastensperre aktivieren / deaktivieren
- 11 4.6.6 Werkseinstellungen wieder herstellen (Reset)
- 13 4.7 Schaltausgänge
- 13 4.7.1 Schaltausgang x: Definition der Schaltcharakteristik
- 14 4.7.2 Schaltausgang x: Obere Schaltgrenze (Einschaltpunkt)
- 15 4.7.3 Schaltausgang x: Untere Schaltgrenze (Rückschaltpunkt)
- 15 4.7.4 Schaltausgang x: Einschaltverzögerung
- 15 4.7.5 Schaltausgang x: Rückschaltverzögerung
- 16 4.7.6 Schaltausgang x: Testen des Schaltausgangs
- 16 4.7.7 Anzeigefunktion der Status-LED ändern
- 18 4.8 Analogausgänge
- 18 4.8.1 Analogausgang x: Zuweisung der oberen Grenze
- 18 4.8.2 Analogausgang x: Zuweisung der unteren Grenze
- 18 4.8.3 Analogausgang x: Festlegen der Signalform
- 19 4.8.4 Analogausgang x: Testen des Analogausgangs
- 19 4.9 Diagnosemöglichkeiten
- 20 4.9.1 Logbuch aufrufen
- 20 4.9.2 Maximale und minimale Temperatur
- 21 4.9.3 Zu protokollierenden Schaltausgang festlegen
- 21 4.9.4 Verzögerung zur Min/Max-Speicherung der Temperatur
- 22 5 Instandhaltung
- 22 5.1 Inspektion und Wartung
- 22 5.2 Service und Reparatur
- 22 5.3 Fehlerbehebung
- 22 6 Entsorgung
- 22 7 Beigefügte Dokumente
- 42 8 Technische Daten
- 43 8.1 Abmessungen
- 44 8.2 Steckerbelegung
- 46 8.3 Aktuelle Einstellungen
- 48 9 Anhang: Übersicht Menüabfolge
- 1 Installation- and Operation Instruction
- 23 1 Introduction
- 24 2 Important advice
- 25 3 Installation and Connection
- 26 4 Operation
- 26 4.1 Switching on
- 26 4.2 LED-Status Display
- 27 4.3 Common Key Functions
- 28 4.4 Key Lock active
- 28 4.5 Overview Menu Structure
- 29 4.6 Changing general Settings
- 29 4.6.1 Setting Unit for Temperature Measurement
- 29 4.6.2 Setting Cable Length
- 29 4.6.3 Setting Cable Cross Section
- 29 4.6.4 Setting the Display Refresh Rate
- 30 4.6.5 Activating / deactivating Key Lock
- 30 4.6.6 Reset Factory Settings
- 32 4.7 Switching Outputs
- 32 4.7.1 Switching Output 1: Definition of the Switching Characteristics
- 33 4.7.2 Switching Output x: Upper Limit (Set Point)
- 34 4.7.3 Switching output x: Lower Limit (Reset Point)
- 34 4.7.4 Switching Output x: Delay for Set Point
- 34 4.7.5 Switching Output x: Delay for Reset Point
- 35 4.7.6 Switching Output x: Testing the switching output
- 35 4.7.7 Switching Output x: Changing Switching Function of LED
- 37 4.8 Analog Outputs
- 37 4.8.1 Analog Output x: Setting the upper Limit
- 37 4.8.2 Analog Output x: Setting the lower Limit
- 37 4.8.3 Analog Output x: Setting the Output Characteristic
- 38 4.8.4 Analog Output x: Testing the Analog Output
- 38 4.9 Diagnostic Tools
- 39 4.9.1 Open the Journal
- 39 4.9.2 Maximum and minimum Temperature
- 40 4.9.3 Assigning the Switching Output for recording
- 40 4.9.4 Delay for recording min/max Temperature
- 41 5 Maintenance
- 41 5.1 Inspection and Service
- 41 5.2 Service and Repair
- 41 5.3 Troubleshooting
- 41 6 Disposal
- 41 7 Attached Documents
- 42 8 Technical Data
- 43 8.1 Dimensions
- 44 8.2 Pin Assignment
- 46 8.3 Actual settings
- 50 9 Appendix: Overview Menu Structure