EL-GFX-DTC Graphing Thermocouple Temperature Data Logger


Add to my manuals
9 Pages

advertisement

EL-GFX-DTC Graphing Thermocouple Temperature Data Logger | Manualzz

EL-GFX-DTC Dual Channel Graphing Thermocouple Temperature Data Logger

Features

• Dual channel, compatible with K, J, and T Thermocouples

• Stand-alone operation

Powered by replaceable batteries

Built-in date and time clock

Deep on-board memory of 252,928 readings

Built-in graphics display for temperature trending, high/ low capture, and more

• IP67 construction resists dust ingress and water submersion up to 1 meter

• Programmable high/low alarm thresholds

• Both audible and visual alarm states

• Two K-type thermocouple probes included (0 to 400°C range)

Total Record Times

Sampling Interval

1 sample every 2 seconds

Record Time

5.8 days

1 sample every 5 seconds

1 sample every 10 seconds

1 sample every 15 seconds

1 sample every 20 seconds

1 sample every 30 seconds

1 sample every 1 minutes

1 sample every 2 minutes

1 sample every 5 minutes

1 sample every 10 minutes

1 sample every 15 minutes

1 sample every 20 minutes

1 sample every 30 minutes

1 sample every 1 hour

14 days

29 days

43 days

58 days

87 days

174 days

249 days

> 2 years

> 2 years

> 2 years

> 2 years

> 2 years

> 2 years

Programmable Elements

• Logger name and channel names

• °F or °C

• Thermocouple type per channel

• Sampling interval

2, 5, 10, 15, 20, 30 seconds

1, 2, 5, 10, 15, 20, 30 minutes

1 hour

• LCD control to maximize battery life

LCD and backlight on for 15 sec after button press

LCD always on, backlight on for 15 sec after button press

• Security setting

Allow/disallow the viewing and the adjustment of settings when deployed

• High and low temperature alarm thresholds

• Alarm delay – the number of consecutive samples after a threshold breech before an alarm is triggered

• Enable/disable status LEDS (sampling, alarm)

• Enable/disable audible alarm

• Record trigger

Immediate

When the logger button is pressed

Less or greater than a programmable temperature value

At a specific date and time

DATAQ Instruments, Inc. • 241 Springside Drive • Akron, Ohio 44333 • Tel: 330-668-1444 • Email: [email protected] • www.dataq.com

EL-GFX-DTC Description

The EL-GFX-DTC data logger offers a cost-effective approach to thermocouple temperature data logging for one or two channels. Its integrated USB interface allows the instrument to be configured by any Windows-based PC. Typical programmable parameters include a unique data logger name and a name for each channel, °F or °C temperature scale, sampling interval (2 seconds to 1 hour), high and low alarm limits, security settings, triggered recording parameters, and more. The GFX-DTC is powered by two replaceable (nonrechargeable) ½ AA lithium batteries, for long-life recordings typically up to one year.

Model EL-GFX-DTC provides a security option that can prohibit tampering of the logger in any manner that would affect recorded data after deployment.

The GFX also supports a programmable alarm delay feature, from immediate to after 250 readings. This is useful where temperature is known to fluctuate momentarily into an alarm threshold, but does not necessarily indicate an alarm condition. For example, the logger is placed inside a freezer or refrigerator. If the door opens the temperature will momentarily rise, but the logger can be configured to ignore that increase for a programmed number of samples before triggering an alarm.

During deployment, status LEDS provide continuous visual feedback of sampling events, battery status, and alarm states. In addition, three front panel function buttons allow you to view the memory consumed (0-100 %), the number of readings contained in the logger’s memory, and the current battery capacity (0-100 %.) The last recorded temperature reading can also be viewed, along with historical data in the form of minimum and maximum values logged during the current recording cycle, as well as an historical trend plot. Using front panel buttons you can also start and stop the logger and view pre-programmed alarm thresholds.

The GFX-DTC is programmed using software that is provided with the instrument. The software programs all operating parameters and options. When a logging session is complete, the GFX logger is plugged into any computer onto which the software is installed using the supplied cable. Options in the software allow the GFX logger to upload time, date and recorded data to the PC where they can be graphed in a cursor-based chart that displays temperature as a function of recorded time and date. A single mouse click can export the data to Microsoft Excel for detailed analysis and report generation.

The EL-GFX-DTC is provided with everything you need to begin taking thermocouple temperature measurements: USB cable for connecting it to a PC, two K-type thermocouple probes (0 to 400°C range), batteries, and software (downloadable, no CD provided.)

Also provided is a convenient wall clip that can be either permanently attached to a surface, or temporarily attached to a ferrous surface via the integrated magnet.

EL-GFX-DTC Close-up

Sampling LED

Status LED

Sensor Window

USB port

Bottom View

Slots for mounting clip

Screw for battery access

Thermocouple

Connections

Function Buttons (3)

330-668-1444 2 www.dataq.com

EL-GFX-DTC Operating Modes

(actual GFX screen images)

Start options (Immediate start not shown):

Delayed manual start

Data display options:

Delayed time/day start Delayed temperature start

Current temperature

(low alarm triggered)

Trend plot with current temp

(high alarm triggered)

General status and local control:

Summary

Logger status screen General control

Locked mode prohibits local adjustments

www.dataq.com 3 330-668-1444

EL-GFX-DTC Specifications

Specification

Measurement Range (K-type TC)

Measurement Range (J-type TC)

Measurement Range (T-type TC)

Internal Resolution

Accuracy (overall error)

Logging Rate (configurable)

Operating Temperature Range***

Battery Life †

Minimum

-200 (-328)

-200 (-328)

-200 (-328) every 2 seconds

-10 (-14)

* Above 999.9°C/F display becomes 1°C. Internal resolution remains at 0.1°C/F.

** At 25°C. See internal accuracy curve (shown right). Important - quoted accuracy is for the

data logger only when measuring within the specified operating temperature.

Thermocouple error is not included and should also be taken into consideration.

*** Operating temperature applies to the data logger module only. Please consult the probe

manufacturer for operating temperature of thermocouple.

† At 25°C and 10 minute logging rate with no alarm LEDs or sounder and minimal LCD use.

Typical

0.1 (0.1)*

±1.5 (±2.7)**

1

Maximum

+1350 (+2462)

+1190 (+2174)

+390 (+734) every 1 hour

+40 (+104)

WARNING: Do not exceed operating temperatures.

Unit

°C (°F)

°C (°F)

°C (°F)

°C (°F)

°C (°F)

°C (°F)

Year

Included

EL-GFX-DTC Ordering Information

Description

EL-GFX-DTC Data Logger

Includes data logger, two (2) batteries (preinstalled), USB cable (31-inches), two (2) 1m K-type

Thermocouple probes (0 to 400°C range), mounting clip, downloadable configuration, review and

Excel export software.

Optional Accessories

Spare Battery

Please note: 2 batteries are required to run the EL-GFX-DTC data logger.

Order Number

EL-GFX-DTC

BAT-3V6

Mounting

Clip

Software

(Via Download)

2 Batteries

(preinstalled)

EL-GFX-DTC

Data Logger

USB Cable

(31 inches)

2 K-Type TC Probes 0 to 400°C Range (1m)

330-668-1444 4 www.dataq.com

EL-GFX-DTC Included Software Overview

Software included with EL-GFX Series instruments allows complete instrument configuration, data retrieval, review, and single-click export facility to Microsoft Excel.

Top EL-GFX Software Menu

Typical EL-GFX Configuration Procedure:

Name the GFX logger and set its temperature scale and logging interval.

Configure the logger’s LCD behavior and security level.

www.dataq.com 5 330-668-1444

EL-GFX-DTC Included Software Overview

(continued)

Configure alarm delay time and methods Configure alarm thresholds and behavior

Configure logging start conditions

Upload Logger Data

Configuration Complete!

Review the stored logging sessions and define where they’ll be saved.

330-668-1444 6 www.dataq.com

EL-GFX-DTC Included Software Overview

(continued)

Review Uploaded Logger Data

Cursor-based review of recorded data with time and date stamps Single-click export to Microsoft Excel.

www.dataq.com 7 330-668-1444

EL-GFX Series Graphing Data Loggers

EL-GFX Data Logger series products are a line of low cost, compact, battery-operated data loggers with a graphical display. Each product in the line offers a specific measurement function and range.

EL-GFX-1 Temperature Data Logger

Measures -30 to +80 °C (-22 to 176 °F)

EL-GFX-DTP Temperature Data Logger with Probe

Measures -40 to +125°C (-40 to +257°F)

EL-GFX-2 Temperature and Humidity Data Logger

Measures -30 to +80 °C (-22 to 176 °F) and 0-100%RH

EL-GFX-2+High Accuracy Temp and RH Data Logger

Measures -30 to +80 °C (-22 to 176 °F) and 0-100%RH

EL-GFX-DTC Thermocouple Temperature Data Logger

Supports K, J, and T type Thermocouples

EL-GFX-TC Thermocouple Temperature Data Logger

Supports K, J, and T type Thermocouples

330-668-1444 8 www.dataq.com

Lascar Data Logger Product Lines

EL-USB Data Loggers EL-GFX Data Loggers EL-WiFi Data Loggers

Lascar EasyLog model EL-USB series products are a line of low cost, compact, battery-operated data loggers with built-in memory and a USB interface for easy setup and data download. Each product in the line offers a specific measurement function (including temperature, voltage, process current, and more).

The EL-GFX line of data loggers is the latest release from Lascar Electronics with similar functionality of the EL-USB series with an added graphic display for data.

Lascar’s EL-Wifi Data Logger series products are low cost, compact, battery-operated data loggers with wireless connectivity to any PC over a WiFi router.

Each product in the line features a large, easy-to-read display of current measurements, and is purchased for specific measurement functions.

241 Springside Drive

Akron, Ohio 44333

Phone: 330-668-1444

Fax: 330-666-5434

Data Acquisition Product Links

(click on text to jump to page)

Data Acquisition | Data Logger

DATAQ, the DATAQ logo and WinDaq are registered trademarks of DATAQ Instruments, Inc. All rights reserves. Copyright © DATAQ Instruments, Inc.

The information on this data sheet is subject to change without notice.

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals