advertisement
EL-GFX-DTC Dual Channel Graphing Thermocouple Temperature Data Logger
Features
• Dual channel, compatible with K, J, and T Thermocouples
• Stand-alone operation
Powered by replaceable batteries
Built-in date and time clock
Deep on-board memory of 252,928 readings
Built-in graphics display for temperature trending, high/ low capture, and more
• IP67 construction resists dust ingress and water submersion up to 1 meter
• Programmable high/low alarm thresholds
• Both audible and visual alarm states
• Two K-type thermocouple probes included (0 to 400°C range)
Total Record Times
Sampling Interval
1 sample every 2 seconds
Record Time
5.8 days
1 sample every 5 seconds
1 sample every 10 seconds
1 sample every 15 seconds
1 sample every 20 seconds
1 sample every 30 seconds
1 sample every 1 minutes
1 sample every 2 minutes
1 sample every 5 minutes
1 sample every 10 minutes
1 sample every 15 minutes
1 sample every 20 minutes
1 sample every 30 minutes
1 sample every 1 hour
14 days
29 days
43 days
58 days
87 days
174 days
249 days
> 2 years
> 2 years
> 2 years
> 2 years
> 2 years
> 2 years
Programmable Elements
• Logger name and channel names
• °F or °C
• Thermocouple type per channel
• Sampling interval
2, 5, 10, 15, 20, 30 seconds
1, 2, 5, 10, 15, 20, 30 minutes
1 hour
• LCD control to maximize battery life
LCD and backlight on for 15 sec after button press
LCD always on, backlight on for 15 sec after button press
• Security setting
Allow/disallow the viewing and the adjustment of settings when deployed
• High and low temperature alarm thresholds
• Alarm delay – the number of consecutive samples after a threshold breech before an alarm is triggered
• Enable/disable status LEDS (sampling, alarm)
• Enable/disable audible alarm
• Record trigger
Immediate
When the logger button is pressed
Less or greater than a programmable temperature value
At a specific date and time
DATAQ Instruments, Inc. • 241 Springside Drive • Akron, Ohio 44333 • Tel: 330-668-1444 • Email: [email protected] • www.dataq.com
EL-GFX-DTC Description
The EL-GFX-DTC data logger offers a cost-effective approach to thermocouple temperature data logging for one or two channels. Its integrated USB interface allows the instrument to be configured by any Windows-based PC. Typical programmable parameters include a unique data logger name and a name for each channel, °F or °C temperature scale, sampling interval (2 seconds to 1 hour), high and low alarm limits, security settings, triggered recording parameters, and more. The GFX-DTC is powered by two replaceable (nonrechargeable) ½ AA lithium batteries, for long-life recordings typically up to one year.
Model EL-GFX-DTC provides a security option that can prohibit tampering of the logger in any manner that would affect recorded data after deployment.
The GFX also supports a programmable alarm delay feature, from immediate to after 250 readings. This is useful where temperature is known to fluctuate momentarily into an alarm threshold, but does not necessarily indicate an alarm condition. For example, the logger is placed inside a freezer or refrigerator. If the door opens the temperature will momentarily rise, but the logger can be configured to ignore that increase for a programmed number of samples before triggering an alarm.
During deployment, status LEDS provide continuous visual feedback of sampling events, battery status, and alarm states. In addition, three front panel function buttons allow you to view the memory consumed (0-100 %), the number of readings contained in the logger’s memory, and the current battery capacity (0-100 %.) The last recorded temperature reading can also be viewed, along with historical data in the form of minimum and maximum values logged during the current recording cycle, as well as an historical trend plot. Using front panel buttons you can also start and stop the logger and view pre-programmed alarm thresholds.
The GFX-DTC is programmed using software that is provided with the instrument. The software programs all operating parameters and options. When a logging session is complete, the GFX logger is plugged into any computer onto which the software is installed using the supplied cable. Options in the software allow the GFX logger to upload time, date and recorded data to the PC where they can be graphed in a cursor-based chart that displays temperature as a function of recorded time and date. A single mouse click can export the data to Microsoft Excel for detailed analysis and report generation.
The EL-GFX-DTC is provided with everything you need to begin taking thermocouple temperature measurements: USB cable for connecting it to a PC, two K-type thermocouple probes (0 to 400°C range), batteries, and software (downloadable, no CD provided.)
Also provided is a convenient wall clip that can be either permanently attached to a surface, or temporarily attached to a ferrous surface via the integrated magnet.
EL-GFX-DTC Close-up
Sampling LED
Status LED
Sensor Window
USB port
Bottom View
Slots for mounting clip
Screw for battery access
Thermocouple
Connections
Function Buttons (3)
330-668-1444 2 www.dataq.com
EL-GFX-DTC Operating Modes
(actual GFX screen images)
Start options (Immediate start not shown):
Delayed manual start
Data display options:
Delayed time/day start Delayed temperature start
Current temperature
(low alarm triggered)
Trend plot with current temp
(high alarm triggered)
General status and local control:
Summary
Logger status screen General control
Locked mode prohibits local adjustments
www.dataq.com 3 330-668-1444
EL-GFX-DTC Specifications
Specification
Measurement Range (K-type TC)
Measurement Range (J-type TC)
Measurement Range (T-type TC)
Internal Resolution
Accuracy (overall error)
Logging Rate (configurable)
Operating Temperature Range***
Battery Life †
Minimum
-200 (-328)
-200 (-328)
-200 (-328) every 2 seconds
-10 (-14)
* Above 999.9°C/F display becomes 1°C. Internal resolution remains at 0.1°C/F.
** At 25°C. See internal accuracy curve (shown right). Important - quoted accuracy is for the
data logger only when measuring within the specified operating temperature.
Thermocouple error is not included and should also be taken into consideration.
*** Operating temperature applies to the data logger module only. Please consult the probe
manufacturer for operating temperature of thermocouple.
† At 25°C and 10 minute logging rate with no alarm LEDs or sounder and minimal LCD use.
Typical
0.1 (0.1)*
±1.5 (±2.7)**
1
Maximum
+1350 (+2462)
+1190 (+2174)
+390 (+734) every 1 hour
+40 (+104)
WARNING: Do not exceed operating temperatures.
Unit
°C (°F)
°C (°F)
°C (°F)
°C (°F)
°C (°F)
°C (°F)
Year
Included
EL-GFX-DTC Ordering Information
Description
EL-GFX-DTC Data Logger
Includes data logger, two (2) batteries (preinstalled), USB cable (31-inches), two (2) 1m K-type
Thermocouple probes (0 to 400°C range), mounting clip, downloadable configuration, review and
Excel export software.
Optional Accessories
Spare Battery
Please note: 2 batteries are required to run the EL-GFX-DTC data logger.
Order Number
EL-GFX-DTC
BAT-3V6
Mounting
Clip
Software
(Via Download)
2 Batteries
(preinstalled)
EL-GFX-DTC
Data Logger
USB Cable
(31 inches)
2 K-Type TC Probes 0 to 400°C Range (1m)
330-668-1444 4 www.dataq.com
EL-GFX-DTC Included Software Overview
Software included with EL-GFX Series instruments allows complete instrument configuration, data retrieval, review, and single-click export facility to Microsoft Excel.
Top EL-GFX Software Menu
Typical EL-GFX Configuration Procedure:
Name the GFX logger and set its temperature scale and logging interval.
Configure the logger’s LCD behavior and security level.
www.dataq.com 5 330-668-1444
EL-GFX-DTC Included Software Overview
(continued)
Configure alarm delay time and methods Configure alarm thresholds and behavior
Configure logging start conditions
Upload Logger Data
Configuration Complete!
Review the stored logging sessions and define where they’ll be saved.
330-668-1444 6 www.dataq.com
EL-GFX-DTC Included Software Overview
(continued)
Review Uploaded Logger Data
Cursor-based review of recorded data with time and date stamps Single-click export to Microsoft Excel.
www.dataq.com 7 330-668-1444
EL-GFX Series Graphing Data Loggers
EL-GFX Data Logger series products are a line of low cost, compact, battery-operated data loggers with a graphical display. Each product in the line offers a specific measurement function and range.
EL-GFX-1 Temperature Data Logger
Measures -30 to +80 °C (-22 to 176 °F)
EL-GFX-DTP Temperature Data Logger with Probe
Measures -40 to +125°C (-40 to +257°F)
EL-GFX-2 Temperature and Humidity Data Logger
Measures -30 to +80 °C (-22 to 176 °F) and 0-100%RH
EL-GFX-2+High Accuracy Temp and RH Data Logger
Measures -30 to +80 °C (-22 to 176 °F) and 0-100%RH
EL-GFX-DTC Thermocouple Temperature Data Logger
Supports K, J, and T type Thermocouples
EL-GFX-TC Thermocouple Temperature Data Logger
Supports K, J, and T type Thermocouples
330-668-1444 8 www.dataq.com
Lascar Data Logger Product Lines
EL-USB Data Loggers EL-GFX Data Loggers EL-WiFi Data Loggers
Lascar EasyLog model EL-USB series products are a line of low cost, compact, battery-operated data loggers with built-in memory and a USB interface for easy setup and data download. Each product in the line offers a specific measurement function (including temperature, voltage, process current, and more).
The EL-GFX line of data loggers is the latest release from Lascar Electronics with similar functionality of the EL-USB series with an added graphic display for data.
Lascar’s EL-Wifi Data Logger series products are low cost, compact, battery-operated data loggers with wireless connectivity to any PC over a WiFi router.
Each product in the line features a large, easy-to-read display of current measurements, and is purchased for specific measurement functions.
241 Springside Drive
Akron, Ohio 44333
Phone: 330-668-1444
Fax: 330-666-5434
Data Acquisition Product Links
(click on text to jump to page)
Data Acquisition | Data Logger
DATAQ, the DATAQ logo and WinDaq are registered trademarks of DATAQ Instruments, Inc. All rights reserves. Copyright © DATAQ Instruments, Inc.
The information on this data sheet is subject to change without notice.
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project