Stryker InTouch FL27 Critical Care Bed Operations Manual
The InTouch FL27 Critical Care Bed is designed to position patients for procedures, therapy, and recovery in healthcare environments. It also allows for patient transport, provides security, measures and displays patient weight, and allows patients to alert caregivers when they require assistance.
Advertisement
Advertisement
InTouch
®
Critical Care Bed
InTouch
®
Intensivpflegebett
Letto per terapia intensiva InTouch
®
InTouch
®
-bed voor kritieke zorg
Κρεβάτι εντατικής θεραπείας InTouch
®
REF
White letters and transparent outline with rectangle
Model/Modell/Modello/Μοντέλο FL27 (2131/2141)
Version 2.5
Operations Manual
Bedienungsanleitung
Manuale d’uso
Gebruikshandleiding
Εγχειρίδιο λειτουργιών
For parts or technical assistance call:
Ersatzteile oder technische Unterstützung:
Per parti o assistenza tecnica chiamare i seguenti numeri
Voor onderdelen of technische bijstand belt u:
Για εξαρτήματα ή τεχνική βοήθεια καλέστε:
USA/VS/Η.Π.Α.: 1-800-327-0770
Canada/Kanada/Καναδάς: 1-888-233-6888
2012/04 B.2
2131-309-005 REV B www.stryker.com
International Addresses
EUROPE HEADQUARTERS
Stryker SA
Cite-Centre, Grand-Rue 90
P.O. Box 1568
1820 Montreux, Switzerland
Phone: 011-41-21-966-12-01
Fax: 011-41-21-966-12-00
EASTERN EUROPE
Stryker SA - Export Business
Grand-Rue 90
P.O. Box 1567
1820 Montreux, Switzerland
Phone: 011-41-21-966-14-00
Fax: 011-41-21-966-14-01
AUSTRALIA
Stryker Australia
8 Herbert Street
St Leonards NSW 2065
Australia
Phone: 61-2-9467-1000
Fax: 61-2-9467-1010
AUSTRIA
Stryker-Howmedica Osterreich GmbH
Millenium Tower
Handelskai 94-96
120 Wien
Austria
Phone: 011-43-1-240-27-6400
Fax: 011-43-1-240-27-6410
BELGIUM
NV Stryker SA
Ikaros Business Park- Fase III
Ikaroslaan 12
1930 Zaventem, Brussels
Belgium
Phone: 011-32-2-717-9210
Fax: 011-32-2-717-9249
DENMARK
Stryker Denmark
Filial of Stryker AB
Postbox 772
1532 Copenhagen
Denmark
Phone: 011-45-33-93-6099
Fax: 011-45-33-93-2069
UK/IRELAND
Stryker UK Limited
Stryker House
Hambridge Road
Newbury, Berkshire
RG14 5EG, England
Phone: 011-44-1635-556-500
Phone: 011-44-1635-262-400
Fax: 011-44-1635-580-300
FINLAND
Stryker AB, Finland
PO 80 Makelankatu 2
00501 Helsinki
Finland
Phone: 011-35-89-774-4680
Fax: 011-35-89-774-46820
FRANCE
Stryker France S.A.S
ZAC - Avenue de Satolas Green
69330 Pusignan
France
Phone: 011-33-472-45-36-00
Fax: 011-33-472-45-36-99
GERMANY
Stryker Howmedica GmbH
Dr. Homer Stryker Platz 1
47228 Duisburg
Germany
Phone: 011-49-2065-837-0
Fax: 011-49-2065-837-837
GREECE
Stryker Hellas EPE
455 Messogion ave
153 43 Agia Paraskevi
Athens, Greece
Phone: 011-30-2-10-600-32-22
Fax: 011-30-2-10-600-48-12
ITALY
Stryker Italia SrL
Via Ghisalba 15B
00188 Roma
Italy
Phone: 011-39-06-33-05-41
Fax: 011-39-06-33-614-067
MIDDLE EAST / NORTH AFRICA
Stryker Osteonics SA
Twin Towers
11th Floor, Suite 1101 & 1102
P.O. Box 41446
Baniyas Road
Dubai, Deira, UAE
Phone: 011-97-14-222-2842
Fax: 011-97-14-224-7381
NEDERLANDS
Stryker Nederlands BV
(P.O. Box 13, 4180 BA Waardenburg)
4181 CD Waardenburg
The Nederlands
Phone: 011-31-418-569-700
Fax: 011-31-418-569-777
NORWAY
Stryker Norway
Norsk Fillial
Nedre Vollgate 3
0158 Oslo
Norway
Phone: 011-47-22-42-22-44
Fax: 011-47-22-42-22-54
POLAND
Stryker Polska Sp. ZO.O
Kolejowa 5/7
01-217 Warsawa
Poland
Phone: 011-48-22-434-88-50
Fax: 011-48-22-434-88-60
PORTUGAL
Stryker Portugal Produtos
Medicos, LTDA.
Avenida Marechal Gomes da Costa, 35
1800-255 Lisboa
Portugal
Phone: 011-35-1-21-839-49-10
Fax: 011-35-1-21-839-49-19
ROMANIA
Stryker Osteonics Romania S.R.L.
19, Leonida Str.
District 2
7000 Bucharest
Romania
Phone: 011-40-2-12-12-11-22
Fax: 011-40-2-12-12-11-33
SOUTH AFRICA
Stryker Osteonis PTY. LTD.
3 Susan Street
Strydom Park
Johannesburg, 2194 - South Africa
Mailing Address:
P.O. Box 48039
2129 Roosevelt Park
Johannesburg, South Africa
Phone: 011-27-11-791-4644
Fax: 011-27-11-791-4696
SPAIN
Stryker Iberia SL c/Sepulveda n17
28108 Alcobendas
Madrid
Spain
Phone: 011-34-91-728-35-00
Phone: 011-34-91-358-20-44
Fax: 011-34-91-358-07-48
SWEDEN
Stryker AB/Scandinavia
Box 50425
SE-204 14 Malmo
Sweden
Phone: 011-46-40-691-81-00
Fax: 011-46-40-691-81-91
Table of Contents
Symbols and Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-4
Warning / Caution / Note Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-5
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-6
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-6
Brief Product Description and Intended Use Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-6
Product Illustration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-7
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-8
Specifications (Continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-9
Summary of Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-10
Optional Zoom ® Drive System (Model 2141 Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-12
Setup Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-13
Optional XPRT™ Therapy Mattress . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-14
Optional PositionPRO™ Patient Repositioning Mattress Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-14
Applying the linens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-15
Optional Zoom ® Drive System (Model 2141 Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-15
Modes of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-16
Powering the Bed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-16
Modes of Operation with the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-16
Base Operation Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-17
Brake / Drive / Neutral Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-17
Electronic Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-17
Manual Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-17
Litter Operation Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-18
CPR Emergency Release . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-18
Lowering the Fowler During a Power Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-18
Foley Bag Hooks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-18
Patient Restraint Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-18
Nurse Call Usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-19
Auxiliary Power Outlet Usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-19
Head and Footboard Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-19
Positioning the Siderails . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-20
Bed Positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-21
Siderail Operation Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-22
Siderail Motion Control Panel, Nurse Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-22
Bed Positioning Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-23
Siderail Brake/Motion Control Panel, Nurse Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-24
Optional Pendant Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-25
Footboard Operation Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-26
Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-26
System Message Indicator Lights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-27
Main Menu: Patient Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-28
Patient Information, New Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-29
Patient Information, Configure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-29
Main Menu: Bed Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-30
Main Menu: Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-31 www.stryker.com
2131-309-005 REV B 1-1
Table of Contents
Footboard Operation Guide (Continued)
Zero Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-33
Weight History Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-34
Change Weight Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-35
Gain or Loss Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-36
Main Menu: Bed Exit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-38
Main Menu: Support Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-40
Air Mattress Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-40
XPRT™ Mattress Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-40
InTouch® integrated with the XPRT™ Mattress . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-40
PositionPRO™ Mattress Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-41
Interaction Between InTouch® and PositionPRO™ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-41
Specific PositionPRO™ Functionality on the InTouch® . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-41
Main Menu: Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-42
Time Setup Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-43
Smart TV Button (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-43
Language Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-43
Advanced Options Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-44
History Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-45
Nurse Control Backlight Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-47
Maintenance Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-47
Main Menu: Lockouts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-48
Main Menu: iBed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-49
iBed, Awareness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-49
iBed, Protocol Reminder™ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-50
iBed, Protocol Reminder™ (Continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-51
iBed, Head of Bed (HOB) History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-52
iBed, Bed Height History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-52
iBed, Calculator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-53
iBed, Converter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-53
iBed, Translations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-54
iBed, Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-55
iBed, The Braden Scale for Predicting Pressure Sore Risk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-56
iBed, Sound Therapy™ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-57
ZOOM ® Drive Operations (Model 2141) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-58
Using the Zoom ® Drive (Model 2141) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-58
Optional Bed Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-59
Bed Extender - FA64234-XXX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-60
Permanent I .V . Pole - FA64221/FA64238 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-61
Dual Head End Permanent I .V . Pole - FA64202 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-63
Line Management System (Accessory Clip) - FA64210 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-65
Monitor Tray - FA64214 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-66
Upright Oxygen Bottle Holder - FA64187 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-68
Right Fit Oxygen Bottle Holder - FA64203 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-69
Pendant with Motion Control Only - FA64209 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-70
1-2 2131-309-005 REV B www.stryker.com
Table of Contents
Optional Bed Accessories (Continued)
Pendant with Motion Control/Nurse Call - FA64194 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-71
Pendant with Motion Control/Smart TV - FA64195 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-72
Pendant with Motion Control/NC/Smart TV - FA64193 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-73
Pendant Clip - FA64186 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-74
Traction Sleeve - FA64215 through FA64219 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-75
Wall Saver - FA64208 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-76
X-Ray Cassette Holder - FA64205 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-77
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-78
Bed Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-78
Mattress Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-78
Preventative Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-79
Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-79
Recycling Passports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-80
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-96
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-96
To Obtain Parts and Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-96
Service Contract Coverage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-96
Service Contract Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-97
Return Authorization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-97
Damaged Merchandise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-97
International Warranty Clause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-97 www.stryker.com
2131-309-005 REV B 1-3
Symbols and Definitions
Warning/Caution, consult accompanying documentation
Safe Working Load
1.6 SYMBOLS
1.6 SYMBOLS
~ Alternating Current
10A 250V
Fuse Rating for Beds with the 100V~ or 120V~ Electric System
Fuse Rating for Beds with the 200V~ or 220V~ or 240V~ Electric System
6.3A 250V
Type B Equipment: equipment providing a particular degree of protection against electric shock, particularly regarding allowable leakage current and reliability of the protective earth connection .
Class 1 Equipment: equipment in which protection against electric shock does not rely on BASIC
INSULATION only, but which includes an additional safety precaution in that means are provided for the connection of the EQUIPMENT to the protective earth conductor in the fixed wiring of the installation in such a way that ACCESSIBLE METAL PARTS cannot become live in the event of a failure of the BASIC
INSULATION .
Mode of Operation: Continuous
IPX4 Protection from liquid splash
Dangerous Voltage
Protective Earth Terminal
Potential Equalization
Medical Equipment approved by the Canadian Standards Association with Respect to Electric Shock,
Fire, Mechanical and Other Specified Hazards .
In accordance with European Directive 2002/96/EC on Waste Electrical and Electronic Equipment, this symbol indicates that the product must not be disposed of as unsorted municipal waste, but should be collected separately . Refer to your local distributor for return and/or collection systems available in your country .
Model Number
Manufacturer
Return To Table of Contents
1-4 2131-309-005 REV B www.stryker.com
Symbols and Definitions
WARNING / CAUTION / NOTE DEFINITION
The words WARNING, CAUTION, and NOTE carry special meanings and should be carefully reviewed .
WARNING
Alerts the reader about a situation, which if not avoided, could result in death or serious injury . It may also describe potential serious adverse reactions and safety hazards .
CAUTION
Alerts the reader of a potentially hazardous situation, which if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property . This includes special care necessary for the safe and effective use of the device and the care necessary to avoid damage to a device that may occur as a result of use or misuse .
Note
This provides special information to make maintenance easier or important instructions clearer .
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-5
Introduction
INTENDED USE
This manual is designed to assist you with the operation of Stryker InTouch® Critical Care Bed Model FL27 (2131/2141) .
Carefully read this manual thoroughly before using the equipment or beginning maintenance on it . To ensure safe operation of this equipment, it is recommended that methods and procedures be established for educating and training staff on the safe operation of this bed .
This Operations Manual is an integral part of the bed and should be included if the bed is sold or transferred .
BRIEF PRODUCT DESCRIPTION AND INTENDED USE STATEMENT
INTENDED MEDICAL INDICATIONS: InTouch is an AC-powered adjustable hospital bed designed to position patients for procedures, therapy, and recovery in healthcare environment, transport patients between bays and procedural rooms, provide patient security, measure and display patient weight, allow patients to alert caregiver when the patient requires emergency assistance or any assistance to improve comfort levels, and monitor patient position to alert caregiver of a deliberate exit or potential fall . The bed has thirty-nine prerecorded clinical phrases in 24 languages and offers music therapy .
INTENDED PATIENT POPULATION: The intended patient population is acute-care human patients . The safe working load (i .e . the sum of the patient, the mattress, and accessory weight) for InTouch is 550lbs (250kg) . This bed is not intended for use with patients 35 inches or less and/or patients that weigh less than 50 pounds .
INTENDED PART OF THE BODY: InTouch is intended to support a patient . It is intended to be used with a sleep surface .
The Frame can come in contact with human skin but a patient should never be on the frame without a support surface .
INTENDED USER PROFILE: InTouch is intended for use in Acute Care settings . These settings may include critical care, step down, progressive care, med/surg, sub-acute care, and post anesthesia care unit (PACU) or other locations as prescribed . Intended operators are health care professionals (nurses, nurse aids, doctors) which can use all bed operations (e .g ., bed motion functions, nurse call, siderail communications, bed exit, therapy options), patient and bystander, which can use bed motion functions, nurse call & siderail communications, and trained professionals for installation, service, and calibration .
INTENDED CONDITION OF USE AND EXPECTED LIFE: Stryker Medical Bed products are designed for a 10 year expected service life under normal use, conditions, and with appropriate periodic maintenance as described in the maintenance manual for each device . Stryker warrants to the original purchaser that the welds on its Bed products will be free from structural defects for the expected 10 year life of the Bed product as long as the original purchaser owns the products .
The product is intended to be used in a healthcare environment, including hospitals, surgery centers, long term acute care centers, and rehabilitation centers . The intended environmental conditions are 10 to 40 degrees Celsius and
30 to 75% RH . The product is compatible with 35” x 84” surfaces, the facility nurse call system, standard Med-Surg equipment, and the facility infrastructure . InTouch is intended to be used with a 6”-8 .5” sleep surface; asleep surface or overlay greater than 6” that offers therapeutic value may be used with, patient supervision .
Return To Table of Contents
1-6 2131-309-005 REV B www.stryker.com
Introduction
PRODUCT ILLUSTRATION
Headboard
Brake Control
Panel
Nurse Control
Panel
Nurse Call
Siderails
Footboard
Control Panel
InTouch®
Screen
Siderail
Release
Levers
CPR Emergency
Release Pedal
Manual
Backup
ZOOM
®
Drive Wheel
(Model 2141 only)
(Under base not shown)
Casters
Foley Bag
Hooks
Footboard www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-7
Introduction
SPECIFICATIONS
Safe Working Load
Bed Weight
Overall Bed Length
Overall Bed
Width
Note: Safe Working Load indicates the sum of the patient, mattress and accessory weight .
550 lbs
Siderails Up
Siderails Down
750 lbs
90”
42”
40”
Base
Litter
Under Bed Clearance
Patient Surface
• Width
• Length
• Length (with Optional Bed Extender)
5”
35”
84”
90”
Seat Section
• Depth 18 .5”
Foot Section
• Length
• Angle
29”
0° to 50°
Fowler Section
• Length
• Width
• Angle
36”
34” to 35”
0° to 70°
249 kg
340 .2 kg
228,6 cm
106,7 cm
102,9 cm
12,7 cm
88,9 cm
213,4 cm
228,6 cm
47 cm
73,7 cm
0° to 50°
91,4 cm
86,4 cm - 88,9 cm
0° to 70°
Fowler
Bed Lift
System
Scale System
Gatch Section
• Length
• Width
• Angle
Bed Lift Time
Capacity
Accuracy:
• For weight from 100 lb to 550 lb
• For weight from 100 lb to 550 lb
• For weight under 100 lb
18”
34” to 35”
0° to 15°
+/- 2% when in Trendelenburg or Reverse Trendelenburg
+/- 2% when Flat
+/- 2 lb when in Trendelenburg or Reverse Trendelenburg
+/- 2 lb when Flat
45,7 cm
86,4 cm - 88,9 cm
0° to 15°
Cardiac Chair
• Standard Cardiac Chair Position
• Enhanced Cardiac Chair Position
Head: 65°, Seat: 17°, Foot: 30°, Trend: 3°
Head: 70°, Seat: 19°, Foot: 47°, Trend: 3°
36 .5” 92,7 cm Length
Height (high) to top of litter
Height (low) to top of litter
33”
16”
Trendelenburg/Reverse Trendelenburg 12°
83,8 cm
40,6 cm
35 seconds maximum from lowest to highest position
550 lbs 249 kg
• For weight under 100 lb
CPR System Speed to level bed from any position
• Fowler
• Foot and Seat
15 seconds
60 seconds
Drive System Speed (Optional ZOOM ®
• Forward
• Backwards
Drive)
3 .6 mph
2 .5 mph
5 .8 km/h
4 .0 km/h
Return To Table of Contents
1-8 2131-309-005 REV B www.stryker.com
Introduction
SPECIFICATIONS (CONTINUED)
Mattress
Maximum
Current
Consumption
Environmental
Conditions
Electrical
Requirements
Battery
Recommended Mattress Size
• With Bed Extender DM64196
(PositionPRO) (35” x 6” x 4”)
• With Bed Extender DM64197
(XPRT™) (35” x 6” x 5 .5”)
Recommended Air Mattress Size
• With Bed Extender DM64196
(PositionPRO) (35” x 6” x 4”)
• With Bed Extender DM64197
(XPRT™) (35” x 6” x 5 .5”)
Without Optional Auxiliary Outlet
(120VAC Only)
35” x 84” x 6”
35” x 90” x 4”
35” x 90” x 5 .5”
35” x 84” x 6” - 8 .5”
35” x 90” x 4”
35” x 90” x 5 .5”
4 .8 Amps
88,9 cm x 213,4 cm x 15,2 cm
88,9 cm x 228,6 cm x 10,1 cm
88,9 cm x 228,6 cm x 13,9 cm
88,9 cm x 213,4 cm x
15,2 cm - 21,6 cm
88,9 cm x 228,6 cm x 10,1 cm
88,9 cm x 228,6 cm x 13,9 cm
With Optional Auxiliary Outlet(s)
(120VAC Only)
Storage
• Ambient Temperature
• Relative Humidity
• Atmospheric Pressure
9 .8 Amps
-40°F to 158°F
10 to 100%
500 to 1060 hPa
-40°C to 70°C
Operating
• Ambient Temperature
• Relative Humidity
• Atmospheric Pressure
• Scale Accuracy
50°F to 104°F
30 to 75%
700 to 1060 hPa
18 °C to 25 °C
10°C to 40°C
64 °F to 77 °F
All electrical requirements meet CSA
C22 .2 No . 601 .1, UL 60601-1 and IEC
60601-1 .60601-2-38 specifications .
100V~, 50-60Hz, 7 .5A - Two 250V, 10A Fuses
120 ± 10% Vac, 50/60Hz - Two 250V, 10A Fuses
120V~, 50-60Hz, 4 .0A (9 .8A with 120V Optional Auxiliary
Outlet) - Two 250V, 10A Fuses
200V~, 50-60Hz, 3 .2A -Two 250V, 6 .3A Fuses
220V~, 50-60Hz, 2 .9A -Two 250V, 6 .3A Fuses
240V~, 50-60Hz, 2 .7A -Two 250V, 6 .3A Fuses
12 V, 17 .6 Ahr, Sealed Lead-Acid Battery (Part Number QDF9188)
3 .0 V 220mAh Lithium Battery, Size 20mm - Varta Int . CR2032
Coin Cell PC Battery Holder, Size 20mm - MDP Int . BA2032
Stryker reserves the right to change specifications without notice.
Specifications listed are approximate and may vary slightly from unit to unit or by power supply fluctuations .
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-9
Summary of Safety Precautions
Before operating the bed, it is important to read and understand all information in this manual . Carefully read and strictly follow the safety guidelines listed below .
Note
To ensure safe operations of the bed, methods and procedures must be established for educating and training hospital staff on the intrinsic risks associated with the usage of electric beds .
WARNING
• To prevent permanent damage to this unit, the unit must reach room temperature prior to conducting any setup and/or unit operations .
• The Weigh system is intended to assist in the monitoring of the patient’s weight variation . Under no circumstances should its reading be used as sole reference for medical treatment .
• Preventative maintenance should be performed at least once a year to ensure all bed features are functioning properly .
• This bed is not intended for pediatric use, i .e, for any patient measuring 35 inches (88,9 cm) or less .
• This bed is equipped with a hospital grade plug for protection against shock hazard . It must be plugged directly into a properly grounded receptacle . Grounding reliability can be achieved only when a hospital grade receptacle is used .
• Shock Hazard - Improper handling of the power cord may result in damage to the power cord and potential shock hazards . If damage has occurred to the power cord, immediately remove the bed from service and contact the appropriate maintenance personnel . Failure to do so could result in serious injury or death .
• Serious injury can result if caution is not used when operating the bed . Operate the bed only when all people and equipment are clear of the electrical and mechanical systems .
• Always apply the brakes when a patient is on the bed or entering/exiting the bed . Serious injury could result if the bed moves while a patient is getting on or off the bed . Once the brake pedal is engaged, push on the bed to ensure the brakes are securely applied . Do the same test when using the electrical brake .
• To help reduce the number and severity of a potential fall when the patient is unattended, keep the siderails in the fully raised position and the sleep surface horizontal in its lowest position, unless his/her medical condition dictates otherwise . When raising the siderails, be sure that you hear the “click” that signals the locked condition . Pull firmly on the siderail to ensure it is locked into position .
• When the sleep surface sections are articulated, ensure that all the patient’s limbs are within the raised siderails to avoid patient injury .
• When a patient’s condition requires greater safety measures for his/her security, use the lockout controls in the footboard control panel to inhibit the siderail functions or remove any optional pendant control and install protective pads on the siderails .
• Siderails, with or without their padded covers, are not intended to serve as restraint devices to keep patient from exiting the bed . Siderails are designed to keep a patient from inadvertently rolling off the bed . It is the responsibility of the attending medical personnel to determine the degree of restraint necessary to ensure a patient will remain safely in bed . Failure to use the siderails properly could result in serious patient injury .
• To reduce risk of injury, ensure the sleep surface is horizontal and in the lowest position with the siderails fully raised and locked when moving the bed with a patient in it .
• To avoid injury to the patient and/or user, do not attempt to move the bed laterally with the steer mode engaged .
The steer wheel cannot swivel .
• The CPR emergency pedal is for emergency use only . When activating the CPR pedal, all people and equipment must be removed from the area below and around the head, thigh and foot sections of the bed or serious personal injury and/or equipment damage could occur .
• The manual backup brake is for emergency use only . It should not be used for any other situation than an emergency or it might get overused when the time comes to use it for an emergency .
• Possible fire hazard exists when this bed is used with oxygen administering equipment other than nasal, mask type or half bed-length tent type . Unplug the bed power cord from the wall when oxygen administering equipment is used .
When using a half bed-length tent type, ensure that the siderails are outside the oxygen tent and that the tent does not extend below the mattress support level .
Return To Table of Contents
1-10 2131-309-005 REV B www.stryker.com
Summary of Safety Precautions
WARNING (CONTINUED)
• The Bed Exit system (optional) is intended only to aid in the detection of a patient exiting the bed . It is not intended to replace patient monitoring protocol . The Bed Exit system signals when a patient is about to exit the bed . The addition or removal of equipment with the Bed Exit system armed must be done using the “Adding or Removing
Equipment with the System Armed” procedure, otherwise the sensitivity of the system may be affected and the readings of the patient’s movements in the bed be erroneous .
• The Weigh system as well as the Bed Exit system must be adjusted if accessories are added when either system is in function .
• The maximum load capacity of the Monitor Tray is 40 lbs (18kg) .
• The maximum load capacity of the tubular support is 150 lbs (68 kg) .
• The Bed Exit system is not designed to be used with patients weighing less than 50 lbs (23 kg) .
• Do not route the power cord between the mattress and the bed frame .
• Do not attach the power cord to any moving parts of the bed frame .
• The power cord could be pinched and may cause electrical shock if a bed extender is used .
• Do not sit on the bed extender . This may cause the bed to flip .
• Do not steam clean, hose off or ultrasonically clean the bed . Do not immerse any part of the bed . The internal electrical parts may be damaged by exposure to water . Hand wash regularly all surfaces of the bed with warm water and a mild detergent . Wipe cleaned surfaces dry to avoid build-up of cleaning substance . Inspect the mattress after each use . Discontinue use if any cracks or rips, which may allow fluid to enter the mattress, are found in the mattress cover . Failure to properly clean the mattress, or dispose of it if defective, may increase the risk of exposure to pathogenic substances and may bring about diseases to the patient and user .
• Always unplug the bed power cord from the wall socket when servicing or cleaning the bed . When working under the bed with the bed in the high position, always apply the brakes and place blocks under the Bed Lift levers to prevent injury in case the Bed Down switch is accidentally pressed .
• Improperly adjusted restraint straps can cause serious injury to a patient . It is the responsibility of the attending medical personnel to determine proper use of the restraint straps and restraint strap locations . Stryker is not responsible for the type and/or use of restraint straps on any of Stryker’s products .
• Use only hospital grade electric equipment consuming 5A or less with the auxiliary power outlet (optional) . The use of standard electric equipment may bring the current leakage to a level unacceptable for hospital equipment .
• Battery posts, terminals and related accessories contain lead and lead compounds, chemicals known to the State of California to cause cancer and birth defects or other reproductive harm . Wash hands after handling .
• Do not modify the InTouch® bed . Modifying the unit can cause unpredictable operation resulting in injury to the patient or user . Modifying the unit will also void this warranty .
CAUTION
• The weight of the I .V . bags should not exceed 22 lbs (10 kg) .
• The mattress thickness should not exceed 6 inches (15,24 cm) .
• When using a mattress thicker than 6 inches (15,24 cm) or a mattress overlay, extra caution and/or patient supervision may be required to reduce the likelihood of occurence of a patient fall .
• Preventative maintenance should be performed at least once a year to ensure all bed features are functioning properly . Ensure that any bed malfunction is promptly reported to your service personnel for immediate attention .
• When large fluid spills occur in the area of the circuit boards, cables and motors, immediately unplug the bed power cord from the wall socket, remove the patient from the bed and clean up the fluid . Have maintenance completely check the bed . Fluids can have an adverse effect on operational capabilities of any electrical product .
DO NOT put the bed back into service until it is completely dried and has been thoroughly tested for safe operation .
Ensure, among other things, that the plastic components being used as covers for the siderail mechanism arms and the foot end casing are removed and that the parts they cover are thoroughly dried .
• Clean Velcro AFTER EACH USE . Saturate Velcro with disinfectant and allow disinfectant to evaporate . (Appropriate disinfectant for nylon Velcro should be determined by the hospital .) www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-11
Summary of Safety Precautions
WARNING (CONTINUED)
• To avoid damage to the siderail mechanisms, do not move the bed using the raised siderails . Move the bed using the handles integrated to the boards .
• Because individual beds may have different options, footboards should not be moved from one bed to another .
Mixing footboards could result in unpredictable bed operation .
• When servicing use only identical replacement parts provided by Stryker .
• Some cleaning products are corrosive in nature and may cause damage to the product if used improperly . If the products described above are used to clean Stryker patient care equipment, measures must be taken to ensure the beds are wiped with a damp cloth soaked in clean water and thoroughly dried following cleaning . Failure to properly rinse and dry the beds will leave a corrosive residue on the surface of the bed, possibly causing premature corrosion of critical components . Failure to follow the above directions when using these types of cleaners may void this product’s warranty .
Note
• Throughout this Operations Manual, the words “right” and “left” refer to the right and left sides of a patient lying on his/her back on the bed .
• The addition of accessories affects the motion of the bed .
• The iBed option is only used as an informational feature and is not supposed to replace normal patient monitoring protocol .
OPTIONAL ZOOM
®
DRIVE SYSTEM (MODEL 2141 ONLY)
In addition to the previous warnings and cautions, all of the following warnings and cautions apply to units equipped with the ZOOM ® .
WARNING
• The 2141 InTouch® Bed is intended for use by trained hospital personnel only . Failure to properly train personnel could result in injury .
• USE CAUTION while maneuvering the unit with the drive wheel activated . Always ensure there are no obstacles near the unit while the drive wheel is activated . Injury to the patient, user or bystanders or damage to the frame or surrounding equipment could occur if the unit collides with an obstacle .
• Use caution when transporting the unit down halls, through doors, in and out of elevators, etc . Damage to the siderails or other parts of the unit could occur if the unit comes in contact with walls or door frames .
• Put the drive wheel in the neutral position and release the brake before pushing the unit manually . For 2141 models; push the Brake Off button to disengage drive wheel (ZOOM ® ) before pushing the unit manually . Do not attempt to push the unit manually with the drive wheel engaged . The unit will be difficult to push and injury could result .
• If unanticipated motion occurs, unplug the power cord from the wall socket, push the battery power on/off switch to the “OFF” (0) position (the LED will not be illuminated), actuate the drive wheel pedal to the neutral position and call maintenance .
• The power save mode is activated after one hour on battery power with no motion release switch activation .
Functions including Bed Exit, Scale and Motion will cease to operate when the unit enters the power save mode .
Injury to the patient could occur if proper patient monitoring protocol is not observed .
• Always unplug the power cord and push the battery power on/off switch to the “OFF” (0) position before service or cleaning . When working under the frame, always support the litter frame to prevent injury in case the Bed Down switch is accidently activated .
Return To Table of Contents
1-12 2131-309-005 REV B www.stryker.com
Setup Procedures
To prevent permanent damage to this unit, the unit must reach room temperature prior to conducting any setup and/or unit operations.
It is important to ensure that the bed is working properly before it is put into service. The following list will help ensure that each part of the bed is checked.
• Install the foot and headboards on the bed . Insert the footboard carefully so that the footboard fit in smoothly .
CAUTION
Because individual beds may have different options, footboards should not be moved from one bed to another . Mixing footboards could result in unpredictable bed operation .
• Plug the power cord to the bed connector at the head end of the bed and into a properly grounded hospital grade wall socket .
• Turn the battery power switch to the “ON” (1) position (located below the litter surface at patient right side) .
• Set the time and date through the touch screen (see page 42 for procedure) .
WARNING
This bed is equipped with a hospital grade plug for protection against shock hazard . It must be plugged directly into a properly grounded receptacle . Grounding reliability can be achieved only when a hospital grade receptacle is used .
• To test the manual backup brake (located on the patients right side only), flip the manual backup brake pedal outward and depress down fully on the foot end side of the pedal . The word BRAKE in red will be visible in the clear window located on the base next to the pedal and the Brake symbol will be flashing on the footboard control panel . The brakes should now be applied .
• To test the electric brake; engage the brake by pressing the brake button on the siderail or footboard control panel .
Try moving the bed to ensure the brakes are applied . Press the Neutral or Brake button on the siderail control panel or on the footboard to disengage the brake .
• For Model 2141 Beds only: On the patient’s right side of the bed, flip the manual backup brake pedal outward and depress down fully on the head side of the pedal . The letter D in GREEN should be visible in the clear window located on the base next to the pedal . The ZOOM ® Drive wheel should now be engaged . Depress down fully on the center of the pedal until the letter N in BLUE is visible in the clear window . The ZOOM ® Drive wheel should now be disengaged .
• Ensure that the siderails raise, lock in the fully raised position and lower smoothly .
• Run through each control on the footboard .
• Verify the scale system and the Bed Exit system for proper operation .
• Run through each control on both inner and outer control panels of the head siderails . If the bed is equipped with the optional Communication Pendant, plug the provided cable to the 37-pin connector located at the head end of the bed and into the proper wall socket .
• Raise the bed completely and activate the Trendelenburg function . Ensure the head end lowers to the full down position . See “Nurse Control” illustration on page 20 . Level the bed using the Bed Lift system controls .
• Raise the bed completely and activate the reverse Trendelenburg function . Ensure the foot end lowers to the full down position . See “Nurse Control” illustration on page 20 . Level the bed using the Bed Lift system controls .
• Verify functionality of the CPR emergency release: raise the fowler up then depress down on the CPR pedal located on either side of the bed at the head end of the bese . The fowler will lower towards a flat position until the pedal is released .
• Ensure the Knee Gatch (if raised) also starts flattening when the Fowler is completely down . Following the complete lowering of the Fowler, wait approximately 30 seconds (the time for the Fowler control motor to reset) and verify that the motor has indeed reset by raising the Fowler fully up using the “Fowler Up” control .
• Verify the following optional equipment for proper operation: 120V auxiliary outlet, night light, emergency crank operation, etc .
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-13
Setup Procedures
OPTIONAL XPRT™ THERAPY MATTRESS
If your bed is equipped with the XPRT™ Therapy Mattress option, perform the following setup procedures to install the Mattress . For graphic representation of the setup procedures, refer to the XPRT™ Therapy Mattress Operations
Manual .
1 . Place mattress over bed litter with printed logo at head end of the bed .
2 . Fold back foot end section of mattress .
3 . Place foot box on foot end of bed litter .
4 . Connect the two (2) color coded connectors on the foot box to the corresponding color coded connectors on the mattress . Connect black connection on foot box to cable adapter 2950-001-180 and then to the bed frame .
5 . Turn the locking collars clockwise to secure the connections .
6 . Connect the other end of the black connector to the bed (left side of litter at foot end behind the accessory outlet) .
7 . Connect the air line from the mattress to the corresponding fitting on the foot box .
8 . Fasten the two (2) retaining clips to the two (2) D-Rings on the foot box .
9 . Lower the foot section of mattress over the foot box . Attach the mattress to the bed frame using the mattress tiedowns .
10 . Apply linens utilizing the “D” rings for the flat sheet .
11 . To secure linens, to mattress, thread four corners through D-Rings attached to mattress .
12 . To ensure proper therapy, do not pull linens taut . The linens should remain loose and wrinkly on the surface of the mattress .
13 . Plug the mattress power cord into a properly grounded, hospital grade receptacle .
Note
The InTouch® bed will detect when the mattress has been connected as well as what type of mattress it is . This will be shown on the InTouch® screen .
WARNING
• Do not route the power cord between the mattress and the bed frame .
• Do not attach the power cord to any moving parts of the bed frame .
• The power cord could be pinched and may cause electrical shock if a bed extender is used .
OPTIONAL POSITIONPRO™ PATIENT REPOSITIONING MATTRESS OPTION
If your bed is equipped with the Position PRO™ Patient Repositioning Mattress option, perform the following setup procedures to install the mattress . For graphic representation of the setup procedures, refer to the Position PRO™
Mattress Operations Manual .
1 . Place the mattress onto the bed .
2 . Flip the foot section towards the head end .
3 . Place the control box (upside down) into the opening in the foot section .
4 . Connect outer transparent tubes to manifold, matching the color coding .
5 . Connect tilt sensor cables .
a . Align the white dots .
b . Twist clockwise to fasten .
6 . Connect pendant cable .
7 . Connect power cord (4’) and turn the switch to on .
Note: The switch is hidden under the power cord .
8 . Fasten straps over the power cord .
9 . Install the power cord in the two (2) retaining clips .
10 . Fasten the three (3) retaining straps .
11 . Carefully rotate the foot end control box and the mattress into the flat position .
Return To Table of Contents
1-14 2131-309-005 REV B www.stryker.com
Setup Procedures
OPTIONAL POSITIONPRO™ PATIENT REPOSITIONING MATTRESS OPTION (CONTINUED)
CAUTION
Gently lower the foot end section to not damage the control box .
12 . Fasten the retaining straps to secure the mattress to the bed frame (four (4) straps total) .
13 . Connect the power cord to the 110V outlet on the bed .
Applying the linens
1 . Apply the linens using the “D” rings for the flat sheet .
2 . To effectively use the “Turn Assist”, do not pull linens taut . Linens should remain loose and wrinkly on surface of the mattress .
WARNING
Ensure that you have always access to the CPR straps .
OPTIONAL ZOOM
®
DRIVE SYSTEM (MODEL 2141 ONLY)
If your bed is equipped with the Optional ZOOM ® Drive System, run through the preceding setup procedures and continue with the procedures listed below .
• With the battery power switch in the “ON” position and the drive wheel in the neutral position (not touching the floor), ensure the “Engage Drive Wheel” LED on the head end control panel is illuminated .
• Run through the operation of the drive wheel (refer to page 57 ) to ensure it is operating properly .
• On both sides of the bed, flip the manual backup brake pedal outward and depress down fully on the head side of the pedal . The letter D in GREEN should be visible in the clear window located on the base next to the pedal . The
ZOOM® drive (2141 Model only) should now be engaged . Depress down fully on the center of the pedal until the letter N in BLUE is visible in the clear window . The ZOOM ® drive should now be disengaged .
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-15
Modes of Operation
POWERING THE BED
When the bed power switch is turned off or in the event of a power failure, the settings of the lockout controls and the calibration data of the Scale and the Bed Exit (optional) systems are preserved .
Note
The bed is provided with batteries which will charge themselves while the bed is connected to the wall socket . When the “Battery Low” light is on you must connect the bed to make sure the battery will be recharged . Also, make sure the battery disconnect switch is turned on .
WARNING
Shock Hazard - Improper handling of the power cord may result in damage to the power cord and potential shock hazards . If damage has occurred to the power cord, immediately remove the bed from service and contact the appropriate maintenance personnel . Failure to do so could result in serious injury or death .
MODES OF OPERATION WITH THE BATTERY
A power outage to the boards will initiate the battery mode . During the operation in the battery mode, the system will set itself in the following modes:
Sleep Mode:
Alarm Mode before Dead Mode: The Alarm Mode before Dead Mode occurs if, while the system is in the Sleep
Mode, the Bed Exit is activated and if there is no activity in regards to the control panels of the bed for a period of 4 hours or the batteries are weak .
Dead Mode:
This mode occurs as a result of the absence of electricity and no activity has been detected from the control panels for 5 minutes . Also, after 1 hour with no activity when the bed is on DC power, the touch screen turns off and the sleep mode occurs .
Normal Mode:
The Dead Mode is activated if, while the bed is in the Sleep Mode, the Bed Exit is not activated and there is no activity on the control panels for a period of 1 hour or the batteries are weak . This mode is also activated after 15 minutes if, while the system is in the Alarm Mode before Dead Mode, the power is not back and there is no activity on the control panels .
This is activated under the following conditions:
1 . The power is back or if any sign of activity is detected back on the control panels .
2 . The touch screen is no longer in the calibration state .
Power Up Mode: If the power supply is back (that is, the bed is plugged back to the wall socket or the power outage is over), the system will exit the Dead Mode to go to the Power Up
Mode if the batteries no longer have power and need to be recharged . The system then goes to the Initialization Mode which goes back to the Normal Mode .
Note
The time needed to charge both batteries completely is approximately 8 hours .
Return To Table of Contents
1-16 2131-309-005 REV B www.stryker.com
Base Operation Guide
BRAKE / DRIVE / NEUTRAL OPERATION
The bed’s brake system can be used in electronic or manual mode .
Electronic Mode
• To engage one of these systems: press on the “Brake”, “Drive” or “Neutral” button located on the footboard and on each head siderail .
• To disengage the brake: press on one of the following buttons: “Brake”, “Neutral” or “Drive” .
• To disengage the neutral and drive, press on one of the two buttons not selected .
These systems can also be disengaged using the pedal . To do so, simply flip the pedal to one of the systems not in use .
Manual Mode
• To engage the brake, drive or neutral system, flip the pedal on the side of the base to the desired system .
• To disengage these systems, flip the pedal to one of the systems not in use . The electronic mode could also be used to disengage the systems .
WARNING
• Always apply the brakes when a patient is on the bed (except during transport) or entering/exiting the bed . Serious injury could result if the bed moves while a patient is getting on or off the bed . Once the brake pedal is manually engaged, push on the bed to ensure the brakes are securely applied .
• To reduce risk of injury, ensure the sleep surface is horizontal and in the lowest position with the siderails fully raised and locked when moving the bed with a patient in it .
• To avoid injury to the patient and/or user, do not attempt to move the bed laterally with the Drive mode activated; the Drive wheel cannot swivel .
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-17
Litter Operation Guide
CPR EMERGENCY RELEASE
When quick access to the patient is needed and the Fowler, Gatch, or Foot end are raised, the bed can be positioned at 0° by pressing on the CPR pedal . The Fowler, Gatch and Foot end will automatically flatten .
WARNING
The CPR pedal is for emergency use only . When activating the CPR pedal, all people and equipment must be removed from the area below and around the head, thigh and foot sections of the bed or serious personal injury or equipment damage could occur .
Lowering the Fowler During a Power Failure
If there is a power failure and the backup battery is out, the CPR pedal can be used to partially or completely lower the
Fowler . Simply depress the CPR pedal until the desired angle is reached .
FOLEY BAG HOOKS
The four Foley bag hooks are located on both sides of the bed frame under the foot and seat sections . Two are centered and isolated .
Note
The Foley bag hooks move when the Fowler is activated . Fowler motion must be locked out when using these hooks to avoid inadvertent movement of the hooks .
PATIENT RESTRAINT LOCATIONS
The bed has 12 locations on the mattress support for installing patient restraint straps . Ten of them are located on the mattress support edges directly across from each other and the remaining two are located on the top edge of the head section .
WARNING
Improperly adjusted restraint straps can cause serious injury to a patient . It is the responsibility of the attending medical personnel to determine proper use of the restraint straps and restraint strap locations . Stryker is not responsible for the type and/or use of restraint straps on any of Stryker’s products.
CAUTION
Clean Velcro AFTER EACH USE . Saturate Velcro with disinfectant and allow disinfectant to evaporate . (Appropriate disinfectant for nylon Velcro should be determined by the hospital .)
Return To Table of Contents
1-18 2131-309-005 REV B www.stryker.com
Litter Operation Guide
NURSE CALL USAGE
The Nurse Call function allows the patient to communicate through a speakerphone with the nurse station by simply pressing the Nurse Call button on the Patient Control . The speakerphones are located on the inner side of the head end siderails .
The communication between the patient and the nurse station is established the moment a member of the nursing staff responds to the call signal .
Note
The Nurse Call system will automatically send a call signal to the nurse station if the communication between the bed and the nurse station is interrupted following a power failure, the switching off of the bed or the disconnection of the
Nurse Call communication cable .
AUXILIARY POWER OUTLET USAGE
The auxiliary power outlet provides the nursing staff with a convenient power source for peripheral equipment . The outlet is located on the left side at the foot end of the bed . A 5A breaker is integrated to the power outlet .
WARNING
Use only hospital grade electric equipment consuming 5A or less with the auxiliary power outlet (optional) . The use of standard electric equipment may bring the current leakage to a level unacceptable for hospital equipment .
HEAD AND FOOTBOARD OPERATION
Both the footboard and the headboard can be removed or replaced easily . The removal of the headboard allows easy access to the patient’s head .
To Remove Boards
1 . Grasp both ends of the board and lift up .
To Install Boards
1 . For the headboard: Insert the board posts inside the mounting sockets and push down to secure it .
2 . For the footboard: Insert the board slowly in the mounting sockets while ensuring that the board connector properly fits on the connector of the foot end casing .
Note
If the footboard must be removed and the bed electric functions remain accessible through the siderails, ensure that the siderail controls are operational by deactivating any lockout activated before removing the footboard . www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-19
Litter Operation Guide
POSITIONING THE SIDERAILS
The bed siderails are designed to allow their lowering using only one hand . They have one lock position which is the upper position .
WARNING
• Keep siderails in the fully raised position and the sleep surface horizontal in its lowest position when the patient is unattended, unless its medical condition dictates otherwise . When raising the siderails, be sure that you hear the
“click” that signals the locked condition . Pull firmly on the siderail to ensure it is locked into position .
• Siderails, with or without their padded covers, are not intended to serve as restraint devices to keep patient from exiting the bed . Siderails are designed to keep a patient from inadvertently rolling off the bed . It is the responsibility of the attending medical personnel to determine the degree of restraint necessary to ensure a patient will remain safely in bed . Failure to use the siderails properly could result in serious patient injury .
To Lower the Siderails (Figure 1.1)
• Lift the yellow latch (A) using one hand and guide it down with the other hand . (See figure below)
To Raise the Siderails (Figure 1.1)
• Pull on the siderail until it is locked in the fully raised position; a “click” will be heard to indicate it is locked . (See figure below)
A
Return To Table of Contents
1-20
Figure 1.1
2131-309-005 REV B
A www.stryker.com
Litter Operation Guide
BED POSITIONING
This bed features five positions: Flat, Standard Chair, Chair Enhanced, Trendelenburg and Reverse Trendelenburg .
These positions can be achieved by pressing their respective buttons on the “Nurse Control” . (See page 1-22 ) .
OPTIONAL BED EXTENDER
The bed can be expanded using a Bed Extender . For further information see page 1-60 .
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-21
Siderail Operation Guide
SIDERAIL MOTION CONTROL PANEL, NURSE CONTROL
A B C D
E
K
F
G
L
(Patient’s Right Side Shown)
Button Name
A
Chair Position
B
C
Flat Position
Trendelenburg
D
Reverse Trendelenburg
E
Fowler Up
F
Gatch Up
G
Foot Up
H
Fowler Down
I
Gatch Down
J
Foot Down
K
Bed Height Up
L
Bed Height Down
H I J
QDF27-2237
Function
Press on this button to place the bed in the Chair or Enhanced
Chair position .
Press on this button to place the bed in the Flat position .
Press on this button to place the bed in the Trendelenburg position .
Press on this button to place the bed in the Reverse Trendelenburg position .
Press on this button to raise the Fowler section of the bed to the desired angle .
Press on this button to raise the Gatch section of the bed to the desired angle .
Press on this button to raise the Foot section of the bed to the desired angle .
Press on this button to lower the Fowler section of the bed to the desired angle .
Press on this button to lower the Gatch section of the bed to the desired angle .
Press on this button to lower the Foot section of the bed to the desired angle .
Press on this button to raise the height of the bed to the desired level .
Press on this button to lower the height of the bed to the desired level .
Return To Table of Contents
1-22 2131-309-005 REV B www.stryker.com
Siderail Operation Guide
BED POSITIONING DESCRIPTIONS
Note
The addition of accessories affects the motion of the bed .
Flat (0˚)
To activate the Flat Bed position, press on the Flat Bed Position button located on the Nurse Control . This will place the Fowler, Foot and Gatch sections in the flat position .
In the event that the Flat button is pressed when the bed is in
Trendelenburg or Reverse Trendelenburg, the bed will be set into motion such that either the head Bed Lift and the foot
Bed Lift will move in order to achieve the Flat Bed position .
Note
It is possible to move the Gatch, Head and Foot sections at the same time .
Standard Cardiac Chair
To activate the Cardiac Chair position, press on the Chair button located on the Nurse Control . Keep pressing the control until the complete Cardiac Chair position is obtained . The Fowler and Gatch will simultaneously raise to full up position and then the bed will position itself into Reverse Trendelenburg .
Enhanced Cardiac Chair
To activate the Enhanced Cardiac Chair position, press on the Chair button located on the Nurse Control . Keep pressing the control until the complete Cardiac Chair position is obtained . The Fowler and Knee Gatch will simultaneously raise to full up position and then the bed will position itself into Reverse Trendelenburg .
Note
There will be a 1 second stop between each chair position .
Trendelenburg
The Trendelenburg position places the bed Fowler to the low position while the foot section is set to the high position .
This position can be achieved by pressing the Trendelenburg button on the Nurse Control until the desired angle is obtained . The bed height can be lowered while it is in Trendelenburg without changing the angle .
Reverse Trendelenburg
The Reverse Trendelenburg position places the bed Fowler to a high position and the foot section to a low position .
This position is achieved by pressing the Reverse Trendelenburg button on the Nurse Control Panel .
Note
• The maximum angle of inclination that can be achieved during the Trendelenburg and Reverse Trendelenburg position is 15 degrees .
• Upon detection of the I .V . pole, the foot section of the bed can no longer be set into motion .
• A minimum clearance of 20 inches is required to allow the Trendelenburg and Reverse Trendelenburg positions to be executed .
• These positions cannot be executed when one or more locks is activated .
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-23
Siderail Operation Guide
SIDERAIL BRAKE/MOTION CONTROL PANEL, NURSE CONTROL
(Patient’s Right Side Shown)
A
N
Brake Off
G
B F
Drive
Brake Set
Vascular
Position
Brake Not Set
QDF27-2239
C
E D
Button Name
A
B
C
D
E
F
G
Brake Off Activation/
Indicator
Steer Activation/Indicator
Brake Activation/Indicator
Brake Not Set
Brake Set
Vascular Position
HOB 30 Degree Position
Function
Press this button to activate the brake off mode . The button will be illuminated when this mode is activated making it an indicator .
Press this button to activate the steer mode . The button will be illuminated when this mode is activated making it an indicator .
Press this button to activate or deactivate the brake mode . The button will be illuminated when this mode is activated making it an indicator .
This LED will be illuminated when the brake system is not activated .
This LED will be illuminated when the brake system is activated .
Press this button to start a treatment with the bed in the vascular position .
Press this button to start a treatment with the bed in the HOB
30 Degree position .
Note
The brake can also be set manually . See page 1-17 for more information .
Return To Table of Contents
1-24 2131-309-005 REV B www.stryker.com
Optional Pendant Operation Guide
OPTIONAL PENDANT CONTROL
G H
C
B
A
CHANNEL
TV
RADIO
I
J
L
D
I
H
F
G
C
D
E
Button Name
A
B
Channel Up
Channel Down
J
Music Up
Music Down
Room Lights
Reading Light
TV
Radio
Nurse Call
Fowler Up
K
Fowler Down
L
Gatch Up
M
Gatch Down
E F M
(All option pendant shown)
K
QDF27-2073
Function
Press this button to change to a higher numbered channel .
Press this button to change to a lower numbered channel .
Press this button to turn the music up .
Press this button to turn the music down .
Press this button to turn the room lights on and off .
Press this button to turn the reading light on and off .
Press this button to turn the TV on and off .
Press this button to turn the Radio on and off .
Press this button to communicate with the nurse station .
Press this button to raise the Fowler section of the bed to the desired angle .
Press this button to lower the Fowler section of the bed to the desired angle .
Press this button to raise the Gatch section of the bed to the desired angle .
Press this button to lower the Gatch section of the bed to the desired angle .
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-25
CONTROL PANEL
A B
Footboard Operation Guide
C
I H G
I
H
F
G
Button Name
A HOB 30 Degree Activation
B
C
Vascular Activation
Display Touch Screen
Function
Press this button to start a treatment with the bed in the HOB
30 Degree position .
Press this button to start a treatment with the bed in the vascular position .
This displays all the functions selected using the buttons . It also shows error codes, alarms and other necessary information .
Quick return to the Main Menu; see below .
D Main Menu
E
Bed Exit On(Armed)/
Off(Disarmed) Indicator
System Message Indicator
Steer Activation/Indicator
Brake Activation/Indicator
Brake Off
Activation/Indicator
When this LED is illuminated, it means the Bed Exit is Armed
(On) . When it is not illuminated, it means the Bed Exit is
Disarmed (Off) . When the LED is flashing, an alarm condition exists .
These LEDs indicate any state the bed might be in or if there is a need to call maintenance . (See descriptions on page 25 .)
Press this button to activate the Steer system . This button is illuminated when the system is activated .
Pres this button to activate or deactivate the Brake system . This button is illuminated when the system is activated .
Press this button to activate the Brake Off system . This button is illuminated when the system is activated .
Return To Table of Contents
1-26 2131-309-005 REV B
D
E
F www.stryker.com
Footboard Operation Guide
SYSTEM MESSAGE INDICATOR LIGHTS
Bed Unplugged:
Bed Charging:
Brake Not Set:
Battery Low:
Call Maintenance:
Locks Enabled:
This indicator shows that the bed is no longer plugged into a wall socket and that its power comes from the battery .
This indicator shows that the bed is actually plugged into a power outlet and that the battery is recharging itself . It is highly recommended not to unplug the bed as long as this LED is on .
This indicator shows that the brakes are not set, which means that the bed might move if it is pushed or pulled . This LED can be seen as a reminder to the personnel that the brake needs to be set before leaving a patient unattended .
This indicator turns on when the power of the battery is at a low level . When it does turn on, the bed should be plugged in a wall socket to allow the battery to be recharged .
This indicator shows when the bed needs repairs or maintenance in regards to its proper functioning . Most of the time the Call Maintenance indicator will be illuminated when there is a problem with the electronic part of the bed system . In this case the appropriate maintenance personnel should be called .
This indicator shows that a part of the bed (Patient Control Only) is locked or that a total lockout (All Bed Controls) is activated . Refer to page 47 for the Lockouts menu .
Note
• The total lockout mechanism blocks the control panel from the siderails, footboard, pendant, touch screen and head panel (located between the ZOOM ® Option handles) .
• The brake system can be activated during a Total Lockout but it cannot be removed at any moment during a Total
Lockout .
• The Total Lockout does not affect the Nurse Call (optional), Scale System (optional) and Bed Exit Detection
(optional) functions .
• The parameters of the lock mechanism will be saved if there is a power outage and will start from their original state when the power returns .
Touch Screen
The Nurse Interface can be activated by pressing one of the function buttons or by pressing on the screen itself . After
2 minutes of inactivity, the LCD intensity will lower to about 10% of its maximal level .
If a submenu is opened and no further action is taken by the user, the LCD will shut off completely after 4 minutes (4 minutes = 2 minutes at 100% intensity + 2 minutes at 10% intensity) .
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-27
Footboard Operation Guide
MAIN MENU: PATIENT INFORMATION
When the “Main” button is pressed, Figure 3 .0 will be displayed . The following features are available to the operator to configure .
• New Patient: pressing on the “New Patient” button displays Figure 3 .1 and patient information may be entered .
• Configure Screen: pressing “Configure Screen” displays Figure 3 .2 and the additional patient information may be configured .
• Height: pressing on the “Height” button allows the operator to enter the patient’s height
• Braden Scale 1 : pressing the “Braden Scale” button takes the operator directly to The Braden Scale for Predicting
Pressure Sore Risk main menu .
• Patient Note: pressing on the pencil tool icon in the bottom right of the Patient Note window allows the operator to enter a patient note if necessary .
• Unit/Room : pressing on the pencil tool icon in the bottom left of the Patient Information screen allows the operator to enter the unit and room number that the patient is in .
1 Copyright . Barbara Braden and Nancy Bergstrom, 1988 . Reprinted with permission .
Patient Information
Patient ID
Patient Since
Weight
Height
BMI
Braden
Scale
ABC2054
01-31-2007
185 lbs
Done:
6 feet 1”
24.5
11
Oct. 24, 2007
11:03 PM
New Patient
Configure
Screen
2.5
Unit/Rm :
Figure 3.0
Oct. 25,2007 12:00 AM
Return To Table of Contents
1-28 2131-309-005 REV B www.stryker.com
Footboard Operation Guide
MAIN MENU (CONTINUED)
Patient Information, New Patient
When the “New Patient” button is pressed on the Patient Information screen, Figure 3 .1 will be displayed . The operator can enter the new patient
ID by pressing the pencil icon which will bring up the keyboard screen . When the patient name is entered, the operator presses the “Ok” button on the keyboard screen and then “Ok” on the New
Patient screen . The Patient Information screen will be displayed with the new patient ID .
Zero Scale
System
Clear HOB And
Bed Height History
Clear Smart Bed
Monitoring
Disarm Bed Exit
Monitoring
Clear All Patient
Information
Clear All Protocol
Reminders
Figure 3.1
Patient Information, Configure
When the “Configure Screen” button is pressed on the Patient Information screen, Figure 3 .2 will be displayed . The operator may enter the patient’s height manually on this screen . Other options available to the operator include calculating the patient’s BMI, displaying the patient’s ID and displaying the last Braden Scale 1 score .
Configure
Display
Height
6 feet 1”
1 Copyright . Barbara Braden and Nancy
Bergstrom, 1988 . Reprinted with permission .
Display Weight
From Scale
System
Calculate BMI
Display Last
Braden Scale
Score
Display
Patient ID
Figure 3.2
Ok Cancel www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-29
Footboard Operation Guide
MAIN MENU: BED CONTROLS
A
J
A
Bed Controls
20”
Unit/Rm :
B
B
C D
C D
E
Oct. 25,2007 12:00 AM
G
H
F
I
Button Name
A
Bed Height Up/Down
B
Fowler Up/Down
C
Gatch Up/Down
D
Foot Up/Down
E
Trendelenburg
F
Reverse Trendelenburg
Angle Indicator
G
I
H
J
HOB Angle Indicator
Battery Power Indicator
Bed Height Indicator
Function
Press the Up arrow to raise the height of the bed or on the
Down arrow to lower it .
Press the Up arrow to raise the Fowler section to the desired angle and on the Down arrow to lower it to the desired angle .
Press the Up arrow to raise the Gatch section to the desired angle and on the Down arrow to lower it to the desired angle .
Press the Up arrow to raise the Foot section to the desired angle and on the Down arrow to lower it to the desired angle .
Press this button to place the bed in the Reverse Trendelenburg position (Head Down / Foot Up) .
Press this button to place the bed in the Trendelenburg position
(Head Up / Foot Down) .
This indicator shows the angle of Trendelenburg or Reverse
Trendelenburg .
This indicator shows the angle of HOB .
This indicator shows the power left in the battery .
This indicator shows the actual height of the bed .
Note
• When pressed the buttons turn grey .
• The bed height is actually the distance between the floor and the bed support mattress .
Return To Table of Contents
1-30 2131-309-005 REV B www.stryker.com
MAIN MENU: SCALE
Footboard Operation Guide
A
J
I
H
G
F
B
C
D
E
Button Name
A
B
Current Patient Weight
Patient Since
I
C
F
G
D
E
H
Scale Last Zero
Change Weight
Gain or Loss
Zero
Weight History
Store
Lbs/Kg
J Delayed Weight
Function
Shows the current weight of the patient .
This will give information about the new patient .
This will give information about the date and time of the last zero of the scale .
Press this button to change the weight of a patient .
Press this button to start the Gain or Loss reading .
Press this button to zero the scale settings .
Press this button to access the History of the patient’s weight .
Press this button to store the patient’s weight .
Press this button to select the measure wanted (pounds or kilograms) .
Press on this button to start the Delayed Weight . This will give the user the time to add or remove accessories .
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-31
Footboard Operation Guide
MAIN MENU: SCALE (CONTINUED)
Delayed Weight Button
When the Delayed Weight button is pressed on the
Scale screen, Figure 5 .1 will be displayed, at which time a 15 second timer starts .
Note
The delayed time is set 15 seconds .
To calculate patient weight with a delayed start:
1 . Press Delayed Weight button .
2 . The 15 second timer starts .
3 . The operator has 15 seconds to lift up any medical devices or equipment that should not be calculated in the patient’s weight . This includes anything lying on the bed surface or resting on the patient (i .e ., hoses) .
4 . After 10 seconds expires, an audible beep will sound .
5 . For the next 5 seconds, the bed will calculate the weight of the patient lying on the bed . The operator must remain lifting equipment described in step 3 .
6 . After 5 seconds expires, another audible beep will sound and the Delayed Weight screen will be displayed as shown in Figure 5 .2 . The bed has completed its calculation and the equipment being lifted can be repositioned on the bed surface .
Message
Weight History
Start weight with delay
Figure 5.1
118.4 lbs
JUL. 22, 2008
09:22 AM
Cancel
Lbs/Kg
The user can change the delayed weight measuring value to Kg or Lbs by pressing on the “Lb/Kg” button located at the bottom of the Delayed Weight screen as shown in the Scale menu .
Previous Next
Figure 5.2
Cancel
Return To Table of Contents
1-32 2131-309-005 REV B www.stryker.com
Footboard Operation Guide
MAIN MENU: SCALE (CONTINUED)
Message
Zero Button
When the “Zero” button is pressed on the Scale screen, Figure 5 .3 will be displayed .
If “Ok” is pressed, the bed will start zeroing the scale and the “Do not touch bed” message will appear as shown in Figure 5 .4 .
If “Cancel” is pressed, the command will be cancelled . If the bed gets into an auto shutoff at this stage, the Zero request will be cancelled .
Scale Zero ?
Cancel
Figure 5.3
Ok
Message
When the “Do not touch bed” message disappears,
Figure 5 .5 will be displayed .
Do not touch bed
Figure 5.4
Close
Message
When the zeroing of the scale is complete, Figure
5 .5 will be displayed .
Note
The “Zero” function will reset all the values for the different patient’s weights . All the data stored in the touch screen concerning the weight of the patient will be lost .
Zero success !
www.stryker.com
2131-309-005 REV B
Close
Figure 5.5
Return To Table of Contents
1-33
Footboard Operation Guide
MAIN MENU: SCALE (CONTINUED)
Weight History Button
When the “Weight History” button is pressed on the
Scale screen, Figure 5 .6 will be displayed .
Each value indicates the time and date that the weight was taken as well as the weight measured .
The cursor allows the user to go further on each side to see the values not visible . The arrows only move by one value at a time .
The “Lb/Kg” button allows the user to change the measuring value .
The “Store” button allows the user to store the reading that has just been done .
The “Close” button will bring the user back to the
Scale main screen .
Figure 5 .7 appears when the user presses on one of the weights displayed on the graphic in Figure
5 .6 . It shows the weight, the date and the time when this weight was taken .
You can choose to go back to the previous one by pressing on “Prev” or to go to the next one by pressing on “Next” .
Note
The system can keep a maximum of fifty (50) entries in memory . Any new value added after the
50 th entry will have the first value deleted following the “First In, First Out” rule .
Weight History
320 lbs
Lbs/Kg
Unit/Rm :
350 lbs
Reset
Weight History
340 lbs
Store
340 lbs
118.4 lbs
JUL. 22, 2008
09:22 AM
375 lbs
Close
Figure 5.6
Oct. 25,2007 12:00 AM
Previous Next
Figure 5.7
Cancel
Return To Table of Contents
1-34 2131-309-005 REV B www.stryker.com
Footboard Operation Guide
MAIN MENU: SCALE (CONTINUED)
Message
Change Weight Button
When the “Change Patient Weight” button is pressed on the Scale screen, Figure 5 .8 will be displayed . Press
“Ok” to change patient weight .
Change
Patient Weight?
Figure 5.8
Ok Cancel
When Ok has been pressed, the “Do not touch bed” message will be displayed as shown in Figure 5 .9 . When the “Do not touch bed” message disappears Figure 5 .10 will be displayed .
Message
Do not touch bed
Press the up or down arrows as show in Figure 5 .10 to increase or decrease the weight . The weight will increment or decrement by 0 .1 lb or kg . If the arrows are pressed longer than just a touch, the incrementation/ decrementation will change by 1 lb or kg . To save the new weight, press “Ok” .
Figure 5.9
Change Patient Weight
Weight
0.0 lbs
Close www.stryker.com
2131-309-005 REV B
Figure 5.10
Ok Cancel
Return To Table of Contents
1-35
Footboard Operation Guide
MAIN MENU: SCALE (CONTINUED)
Gain or Loss Button
When the “Gain or Loss” button is pressed on the
Scale screen, Figure 5 .11 will be displayed . This will show the initial weight of the patient, the current weight and the gain or loss of weight of the patient .
To change the measuring values, press on the
“Lb/Kg” button .
To close the window, press on the “Close” button .
Gain or Loss
Patient Weight
Initial
Current
Gain
127.1 lbs
127.1 lbs
0.0 lbs lbs/Kg
Reset
Close
Figure 5.11
Message
When the reset button is pressed, Figure 5 .12 will be displayed . If “Ok” is pressed, the message “Do not touch bed” will be displayed .
Reset
Gain or Loss?
Figure 5.12
Ok Cancel
Message
When the “Do not touch bed” message disappears, the “Gain or Loss reset success!” message will be displayed as shown in Figure 5 .13 .
Gain Or Loss reset success!
Return To Table of Contents
1-36 2131-309-005 REV B
Figure 5.13.
Close www.stryker.com
Notes www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-37
MAIN MENU: BED EXIT
A
Bed Exit
Footboard Operation Guide
B C
Arm Disarm
Room #:
Oct. 25,2007 12:00 AM
C
D
E
F
D
Button Name
A Traditional Bed Exit Zone
B Medium Zone
Small Zone
Arm
Disarm
Volume ( see page 38 )
E F
Function
Press this button to activate the complete zone of detection .
This is the default zone .
Press this button to activate the medium zone of detection .
Press this button to activate the small zone of detection .
Press this button to Arm the Bed Exit Detection System .
Press this button to Disarm the Bed Exit Detection System .
Press this button to access the Volume menu for the alarm .
Note
• A minimum of 5 lbs is necessary to arm the Bed Exit System .
• During a power outage, the parameters that were set for the detection of the Bed Exit are maintained and saved .
When power resumes, the system will reactivate itself to the same state it was before the power outage . The zone that was selected before the power outage and the signal CALL NURSE on the wall of the hospital (optional) are activated .
• The blue color on one of the zones indicates that this is the zone which is actually activated .
• When the zone color turns red, it means that there is a Bed Exit Alarm . In that case the Event Manager window will appear .
Return To Table of Contents
1-38 2131-309-005 REV B www.stryker.com
Footboard Operation Guide
MAIN MENU: BED EXIT (CONTINUED)
Alarm settings
Volume
A
C
Tone 1
Test
B
D
E
Close
The Volume button allows the user to set an alarm reminder . The alarm has different tones that can be chosen by the user .
Button Name
A Volume Down
B Volume Up
C
D
E
Change Tone
Change Tone
Test Volume / Tone
Function
Press this button to decrease the volume .
Press this button to increase the volume .
There are 10 tones to choose from, and they are labeled 1 through 10 . Press this button to change the tone sound to a lower number than what it is currently set at .
If the bed is equipped with the iAudio feature, then the last available tones will be voices .
Press this button to change the tone sound to a higher number than what it is currently set at .
Press this button to test the current volume or tone setting .
When the volume or tone is changed from its current setting, you may press the test button to hear the new tone setting or the new volume adjustment .
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-39
Footboard Operation Guide
MAIN MENU: SUPPORT SURFACES
Air Mattress Options
The InTouch® bed can detect if it is connected to a mattress . It can also detect the typical Air Mattress which is connected to it .
When the Support Surface button is pressed on the main menu bar and there is no mattress connected to the InTouch®, Figure 7 .0 will be displayed .
No mattress detected.
Please check cable connections
Figure 7.0
XPRT™ Mattress Specifications
Refer to the XPRT™ manual .
InTouch® integrated with the XPRT™
Mattress
Siderail Position
The position of the siderails determines the functionality of the XPRT™ mattress . All the therapies, including Max Inflate, Firmness and Turn Assist, will function if the siderails are unlocked, except for the rotation therapy . If a siderail is unlocked during the rotation therapy, this therapy will automatically stop .
If two therapies are simultaneously executed and a siderail is unlocked during these therapies, they will automatically stop . If one of the siderails is unlocked during a rotation therapy (ex .: rotation and vibration is being executed simultaneously), both therapies will stop instantly .
Unit/Rm :
Rotation
Percussion
Rotation:
Status
OFF
Vibration
Max Inflate
Firmness
Turn Assist
Percussion:
Vibration:
OFF
OFF
Siderails Must be Up
When Patient is Unattended
Start
Stop
Figure 7.1
Oct. 25,2007 12:00 AM
Return To Table of Contents
1-40 2131-309-005 REV B www.stryker.com
Footboard Operation Guide
MAIN MENU: SUPPORT SURFACES (CONTINUED)
PositionPRO™ Mattress Specifications
Refer to the PositionPRO™ manual .
Interaction Between InTouch® and
PositionPRO™
Siderails Position
The position of the siderails determines the functionality of the PositionPRO™ mattress . All the therapies, including
Max Inflate, Firmness and Turn Assist, will function if the siderails are unlocked . If a siderail is unlocked during the rotation therapy, this therapy will automatically stop . If two therapies are simultaneously executed and a siderail is unlocked during these therapies, they will automatically stop . If one of the siderails is unlocked during a rotation therapy (ex .: rotation and vibration is being executed simultaneously), both therapies will stop instantly .
PositionPRO
TM
Left Turn
Firmness (mmHg)
Timer
Max Inflate
120:00
Minutes
Stop
Right Turn
Alarms
Reset
CPR
CPR
Call
Maintenance
Lock All
14 18 22 26 30
Unit/Rm :
Figure 7.3
Oct. 25,2007 12:00 AM
PositionPRO
TM
Start
Right Turn
Timer
+
120:00 Minutes
-
Matress Turn In Progress
Back
Specific PositionPRO™ Functionality on the
InTouch®
The indicators on the PositionPRO™ menu screen from the InTouch® and on the PositionPRO™ pendant are slightly different, but the same functions are available .
When requesting a Turn Assist left or right on the
PositionPRO™ via the InTouch®’s touch screen, a Turn
Assist menu screen will appear (see Figure 7 .3) .
If the InTouch® senses an error with the PositionPRO™, the Call Maintenance LED will turn on (see Figure 7 .3) .
When activating the InTouch® CPR pedal, the
PositionPRO™ automatically turns on .
When the PositionPRO™ is connected to the InTouch®, the PositionPRO™ uses the fowler and Trendelenburg tilt sensors of the InTouch® to sense the angle of the fowler .
Unit/Rm :
PositionPRO
TM
Start
Left Turn
Figure 7.4
Oct. 25,2007 12:00 AM
Timer
+
120:00 Minutes
-
Matress Turn In Progress
Back
Unit/Rm : Oct. 25,2007 12:00 AM
Figure 7.5
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-41
Footboard Operation Guide
MAIN MENU: OPTIONS
A
H
G
B C
Options
Time Setup Smart TV Language
Advanced
Options
History
Nurse Control
Backlight
Maintenance
Unit/Rm : Oct. 25,2007 12:00 AM
D
E
F
Button Name
A Time Setup
B Smart TV
C
D
E
F
Language
Advanced Options
Maintenance
Battery
G
H
Backlight
History
Function
Press this button to setup the time and date on the bed .
Press this button to access the Smart TV menu .
Press this button to access the Language menu and select the language of the touch screen .
Press this button to access the Advanced Options menu .
Press this button to access the Maintenance menu .
This shows the percentage of power left in the battery .
Each battery block indicates 25 percent of the power .
Press this button to access the Backlight menu .
Press this button to access the History menu .
Return To Table of Contents
1-42 2131-309-005 REV B www.stryker.com
Footboard Operation Guide
MAIN MENU: OPTIONS (CONTINUED)
Time Setup Button
When the “Time Setup” button is pressed on the
Options screen, Figure 8 .0 will be displayed . The time and date can now be changed using the arrows . Press “Accept” when done or “Cancel” to refuse the changes done .
Note
The “Time Setup” does not automatically adjust for daylight savings time .
Time Setup
Set Hour Set Date
+ +
9 H 54 AM
-
+ +
2007 Jun 17
-
Smart TV Button (Optional)
When the “Smart TV” button is pressed on the
Options screen, Figure 8 .1 will be displayed . . The television currently attached to the room interface card of the bed can now be selected . Once the selection is done, press on the “Accept” button to confirm the choice .
Language Button
When the “Language” button is pressed on the
Options screen, Figure 8 .2 will be displayed . You can now select the desired language .
Accept
Figure 8.0
Cancel
Smart TV
No Smart TV
TV Configuration
1
Autoconfig 2
Auto Dig Vol 3 6
4
5
Accept
Figure 8.1
Cancel
Language
English
Español
Français www.stryker.com
2131-309-005 REV B
Cancel
Figure 8.2
Return To Table of Contents
1-43
Footboard Operation Guide
MAIN MENU: OPTIONS (CONTINUED)
Advanced Options Button
When the “Advanced Options” button is pressed on the
Options screen, Figure 8 .3 will be displayed .
The operator may activate or deactivate the ZOOM®
Voice , Awareness and or Reminder . If the ZOOM®
Voice option is activated, sound will be played through the speakers when the ZOOM® is activated . If the
Awareness and/or Reminder options are selected, the alarms will be sent through the bed Nurse Call signal .
Figure 8.3
Return To Table of Contents
1-44 2131-309-005 REV B www.stryker.com
Footboard Operation Guide
MAIN MENU: OPTIONS (CONTINUED)
History Button
When the “History” button is pressed on the
Options screen, Figure 8 .5 will be displayed .
When the “Scale” button is pressed, the Weight
History screen appears ( see pg . 32, Fig . 5 .6
) .
History
Scale
Bed Height
Head of Bed
When the “Head of Bed” button is pressed on the Options screen, Figure 8 .6 will be displayed .
The “Head of Bed History” displays the head’s angle of the bed for the last 24 hours .
When the “Angle History” button is pressed, on the Head of Bed screen, Figure 8 .7 will be displayed .
Note
The maximum time frame for angle position is
30 days .
Figure 8.5
Head of Bed
Jul. 22, 2003 09:22 AM
Angle 20°
Duration 02h24min
Close
Angle
History
Figure 8.6
Close
Angle History
Angle
0 ~ 9°
10 ~ 19°
20 ~ 29°
30 ~ 39°
40 ~ 49°
50 ~ 59°
60 ~ 69°
70 ~ 79°
80 ~ 89°
HOB 30
History
Last 12h
12:00
12:00
12:00
12:00
12:00
12:00
12:00
12:00
12:00
Reset
Figure 8.7
Last 24h
24:00
24:00
24:00
24:00
24:00
24:00
24:00
24:00
24:00
Close www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-45
Footboard Operation Guide
MAIN MENU: OPTIONS (CONTINUED)
When the “Bed Height” button is pressed on the
History screen, Figure 8 .8 will be displayed .
Bed Height
Jul. 22, 2003 09:22 AM
Height 28 in
Duration 02h24min
When the “Height history” button is pressed on the
Bed Height screen, Figure 8 .9 will be displayed .
The “Bed Height History” provides the user with information on the current bed height as well as the bed height for the last 24 hours . If the Bed
Height History needs to be reset, press the “Reset” button .
Height history
Figure 8.8
Close
Height history
Height
16 ~ 18
18 ~ 20
20 ~ 22
22 ~ 24
24 ~ 26
26 ~ 28
28 ~ 30
30 ~ 33
Last 12h
00:00
00:00
00:00
00:00
00:00
00:00
00:00
00:00
Low height
History
Reset
Figure 8.9
Last 24h
00:00
00:00
00:00
00:00
00:00
00:00
00:00
00:00
Close
Return To Table of Contents
1-46 2131-309-005 REV B www.stryker.com
Footboard Operation Guide
MAIN MENU: OPTIONS (CONTINUED)
Nurse Control
Nurse Control Backlight Button
When the “Nurse Control Backlight” button is pressed on the Options screen, Figure 8 .10 will be displayed . The backlighting interface allows the user to set the intensity of the indicators (LED) and of the touch screen .
Press the arrows to increase or decrease the intensity . When completed, press the “Close” button .
Maintenance Button
When the “Maintenance” button is pressed on the
Options screen, Figure 8 .11 will be displayed .
When the “Error Codes” button is pressed, Figure
8 .12 will be displayed . The Error codes display provides information on component status .
When the “Signal Values” button is pressed, Figure
8 .13 will be displayed . The Signal Values display provides information on sensor status .
Note
Refer to the Maintenance Manual for further details on Buttons Pressed, Boards and Bed Information .
Close
Figure 8.10
Maintenance
Error Codes
Signal Values
Boards
Buttons Pressed
Bed
Informations
Input
States
Close
Figure 8.11
Error Codes
Current error
ADC invalid values
Battery absent
Battery bad capacity
Cmd without safe from nurse (SRR)
GPIO failure init
Load cell 1 overrange
Motor gatch overheat
Motor brake overload
Motor HL foot overheat
Motor zoom overheat
Safe without cmd from B1
Tilt hilo foot over range
Tilt gatch over range
Brake bad calib.
Brake unable to elec.
Unit/Rm :
Error log
ADC invalid values
Battery absent
Battery bad capacity
Cmd without safe from nurse (SRR)
GPIO failure init
Load cell 1 overrange
Motor gatch overheat
Motor zoom overheat
Safe without cmd from B1
Tilt hilo foot over range
Clear log
Back
Figure 8.12
Oct. 25,2007 12:00 AM www.stryker.com
2131-309-005 REV B
Signal Values (All)
Sensor Values
Temp
24 Vdc
Battery Voltage
VBus
Brake Pot
Charger Amp
31.2 C
26.2 V
25.1 V
24.7 V
70
0.00 A
Load cells values
Load cells
Angle dep. trend
Angle indep. trend
Angle dep. rev. trend
Angle indep. rev. trend
Bed exit
Bed height
Foot end Height
Tilt Values
Head
Gatch
Foot
Base
Hilo foot
Trend
Angle
-13.6
11.6
-10.9
0.0
76.0
11.6
Raw
3980
5211
4334
5246
5200
5411
1
45.9 lb
34062
-27540
32095
-25573
2
57.5 lb
-27139
28195
-25262
26318 x = 54, y = 22
711
869
Zero
5029
5439
5071
5247
4627
5172
3
76.3 lb
-18274
24436
-22896
29058
Gain
813962
-880336
796641
800000
791613
838363
4
100.0lb
24793
-24573
29827
-29607
Close
Unit/Rm :
Figure 8.13
Return To Table of Contents
1-47
MAIN MENU: LOCKOUTS
Lockouts
Footboard Operation Guide
A
B
All bed controls
Patient control only
F
G
Oct. 25,2007 12:00 AM Unit/Rm : i
Bed
Button Name
A
All bed controls
B
Bed Height Lock
C
Fowler Lock
Gatch Lock
D
E
Foot Lock
F
All bed controls
G
Patient controls only
C D E
Function
Press on this button to activate or deactivate the complete lock of the bed .
Press on this button to activate or deactivate the lock for the bed height .
Press on this button to activate or deactivate the lock of the
Fowler .
Press on this button to activate or deactivate the lock of the
Gatch .
Press on this button to activate or deactivate the lock of the
Foot .
Select this when you want to restrain any change in the lock activation on every bed controls .
Select this when you want to restrain any change in the lock activation on the patient controls only .
Note
• If the bed is in Trendelenburg or Reverse Trendelenburg the bed will be locked in this position .
• When one of the locks is enabled, the indicator on the footboard will turn on so the user is aware that some motion may not be activated .
Return To Table of Contents
1-48 2131-309-005 REV B www.stryker.com
Footboard Operation Guide
MAIN MENU: iBED
When the “iBed” button is pressed on the main menu bar, Figure 10 .0 will be displayed .
Note: If your bed is equipped with the iDocumentation option then the iDocumentation button will be displayed on the iBed menu . If your bed is equipped with the
iAudio option then the Sound Therapy™ button will be displayed .
iBed, Awareness
When the “Awareness” button is pressed on the iBed screen, Figure 10 .1 will be displayed .
In this menu the user can monitor some functions or even all of them . To select functions actually in use, press on “Monitor current bed situation” .
When functions are being monitored and an alarm is set the following occurs:
• Footboard LED flashes amber,
• Siderail LED flashes amber,
• Monitored event button turns amber,
• The Event Manager window appears as shown in
Figure 10 .2 .
i
Bed
Awareness
Protocol reminder
HOB History
Conversions Translations
Bed Height
History
Calculator
Braden Scale Documentation
Sound
Therapy
Unit/Rm :
Figure 10.0
Smart bed position
Right head siderail up
Right foot siderail up
Brake on
Bed at low
height
Bed exit monitoring
Monitor current bed situation
Deselect all
Back
Oct. 25,2007 12:00 AM
Left head siderail up
Left foot siderail up
Fowler angle superior to 30
Bed supine
Vascular position
In the Event Manager the user has the choice to:
• PROCEED TO MENU: the user will be directed in the menu where the alarm comes from .
• CLOSE AND DISARM: the alarm will simply be deactivated and no longer monitored .
• REMIND ME IN: the user can decide to report the alarm by choosing the appropriate time reminder .
Note
It is possible to browse through the bed’s menus even if conditions are not corrected . If the touch screen is inactivated for 60 seconds without correcting the conditions, it goes automatically back to the “Smart Bed
Position” screen until the corrections are made .
Unit/Rm :
Figure 10.1
Oct. 25,2007 12:00 AM
Event Manager
Proceed
To Menu
Close And
Disarm
Remind Me In
The following events occured
Bed Exit Alarm Activated
Head Left Siderail Lowered
10 Minutes 20 Minutes 30 Minutes 40 Minutes
Unit/Rm : 12:00 AM
Figure 10.2
Return To Table of Contents
1-49 www.stryker.com
2131-309-005 REV B
Footboard Operation Guide
MAIN MENU: iBED (CONTINUED)
iBed, Protocol Reminder™
When the “Protocol Reminder™” button is pressed on the iBed screen, Figure 10 .3 will be displayed .
To add or view a protocol, press on the appropriate button .
To add a one-time Protocol Reminder™, enter the desired time for the protocol alarm as shown in
Figure 10 .4 .
To add a repeated Protocol Reminder™, enter the lapse of time between each protocol alarm as shown in Figure 10 .5 .
Note
The default screen is the one-time reminder, but the user can change it to “Repeated Task” by pressing on the “Repeated Task” button .
Reminder
Interventions
Pulmonary Skin Care
Fall Prevention Labs / GI
Mobility protocol
* Customizable reminders
Lines /
Dressings
Restraints
Neuro
Custom
View
Reminder’s
List
View
Reminder’s
Log
Back
Unit/Rm :
Figure 10.3
Add A Reminder
Remind At
+ +
9 54 AM
-
Nov 13,
2008
Reminder
Skin Assessment
Repeat Every
+
-
Oct. 25,2007 12:00 AM
Add to reminders list
Repeated
Task
Back
Unit/Rm :
Figure 10.4
Oct. 25,2007 12:00 AM
Add a reminder
Remind At
+ +
9 54 AM
-
Nov 13,
2008
Reminder
Custom
Repeat Every
+
9 Hours
-
Add to reminders list
One time task
Back
Unit/Rm :
Figure 10.5
Oct. 25,2007 12:00 AM
Return To Table of Contents
1-50 2131-309-005 REV B www.stryker.com
Footboard Operation Guide
MAIN MENU: iBED (CONTINUED)
iBed, Protocol Reminder™ (Continued)
When it is time for a reminder, Figure 10 .6 will be displayed .
Reminder alarm!
The following Reminder(s) require Attention
Blood Pressure
Medication
Remind me in
10 Minutes 20 Minutes 30 Minutes
Have done / Will do the necessary action
Did not / Can not do the necessary action
40 Minutes
When the View Reminder’s List button is pressed on the
Reminder screen, Figure 10 .7 will be displayed .
Unit/Rm : Oct. 25,2007 12:00 AM
Figure 10.6
Reminders List
Turning 2008/11/14 - 10:28 AM
Moisture Check 2008/11/17 - 9:20 AM
Unit/Rm :
Edit
Remove from
Reminders
Remove all
Reminders
Back
Figure 10.7
Oct. 25,2007 12:00 AM
Reminders Log
Nutrition Check 2008/11/13 - 10:29 AM
When the View Reminder’s Log button is pressed on the
Reminder screen, Figure 10 .8 will be displayed .
Unit/Rm :
Clear
Log
Back
Figure 10.8
Oct. 25,2007 12:00 AM www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-51
Footboard Operation Guide
MAIN MENU: iBED (CONTINUED)
iBed, Head of Bed (HOB) History
When the “HOB History” button is pressed on the
iBed screen, Figure 10 .9a will be displayed .
The “Head of Bed History” displays the head’s angle of the bed for the last 24 hours .
Head of Bed
Jul. 22, 2003 09:22 AM
Angle 20°
Duration 02h24min
When the “Angle History” button is pressed on the Head of Bed screen, Figure 10 .9b will be displayed .
iBed, Bed Height History
When the “Bed Height History” button is pressed on the iBed screen, Figure 10 .10a will be displayed .
Angle
History
Figure 10.9a
Close
Angle History
Angle
0 ~ 9°
10 ~ 19°
20 ~ 29°
30 ~ 39°
40 ~ 49°
50 ~ 59°
60 ~ 69°
70 ~ 79°
80 ~ 89°
HOB 30
History
Last 12h
12:00
12:00
12:00
12:00
12:00
12:00
12:00
12:00
12:00
Reset
Figure 10.9b
Last 24h
24:00
24:00
24:00
24:00
24:00
24:00
24:00
24:00
24:00
Close
Bed Height
Jul. 22, 2003 09:22 AM
Height 28 in
Duration 02h24min
When the “Height History” button is pressed on the Bed Height screen, Figure 10 .10b will be displayed .
The “Bed Height History” provides the user with information on the current bed height as well as the bed height for the last 24 hours . If the Bed
Height History needs to be reset, press the “Reset” button .
Return To Table of Contents
1-52 2131-309-005 REV B
Height history
Figure 10.10a
Height history
Height
16 ~ 18
18 ~ 20
20 ~ 22
22 ~ 24
24 ~ 26
26 ~ 28
28 ~ 30
30 ~ 33
Last 12h
00:00
00:00
00:00
00:00
00:00
00:00
00:00
00:00
Close
Last 24h
00:00
00:00
00:00
00:00
00:00
00:00
00:00
00:00
Low height
History
Reset
Figure 10.10b
Close www.stryker.com
Footboard Operation Guide
MAIN MENU: iBED (CONTINUED)
iBed, Calculator
When the “Calculator” button is pressed on the iBed screen, Figure 10 .12 will be displayed .
The operator can perform various calculations using the
Calculator tool . If conversions are needed, the operator can switch directly from the Calculator display to a conversion tool by pressing the “Converter” button .
Calculator
7
4
1
0
8
5
2
.
9
6
3
=
+
/
-
+
AC
C
0 Convert
Close
Unit/Rm :
Oct. 25,2007 12:00 AM
Figure 10.12
iBed, Converter
When the “Converter” button is pressed on the iBed screen, Figure 10 .13 will be displayed . When the operator completes their conversions, they may switch back to the “Calculator” by pressing on the “Calculator” button .
Converter
7 8 9
4 5 6
1 2 3
0
.
0
Calcul
Close lbs kg
°C °F mL L
C
°F °C cm in
L cc in cm kg lbs cc L mg oz oz mg L mL
Unit/Rm #:
Oct. 25,2007 12:00 AM
Figure 10.13
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-53
Footboard Operation Guide
MAIN MENU: iBED (CONTINUED)
iBed, Translations
When the “Translation” button is pressed on the
iBed screen, Figure 10 .14 will be displayed .
The Translations display provides information to the operator to retrieve a quick translation of medical sentences in 24 different languages .
The operator may choose a sentence in the list provided on the display and then select the language to be translated in . The translated sentence will be displayed in the Translations box .
Audio is provided for the translations as well .
Press on the play button to hear the translations .
To increase the audio, press the “+” button . To decrease the audio, press the “-” button .
Note: If the iAudio option is not equipped on your bed, then the audio controls buttons will not be dispalyed in Figure 10 .14
Translations
Theme
Neurological Assessment
Pain
Needs Assessment
Awareness & Introduction
Respiratory
Language
Spanish
French
German
Russian
Polish
Italian
Theme: Pain
Do you feel pain?
Where do you feel pain?
What is your pain level?
Are you feeling better?
Translation
Unit/Rm :
Back
Figure 10.14
Oct. 25,2007 12:00 AM
Return To Table of Contents
1-54 2131-309-005 REV B www.stryker.com
Footboard Operation Guide
MAIN MENU: iBED (CONTINUED)
iBed, Documentation
When the “Documentation” button is pressed on the iBed screen, Figure 10 .15 will be displayed .
The operator may retrieve logged patient information for a specific date by selecting the date using the up and down arrows for the year, moth and day . Press the View button and a log will be displayed for that date . You may view specific logged information by pressing either the Bed Exit, Reminder, Scale, iBed or System buttons . www.stryker.com
Figure 10.15
2131-309-005 REV B
Return To Table of Contents
1-55
Footboard Operation Guide
MAIN MENU: iBED (CONTINUED)
iBed, The Braden Scale for Predicting
Pressure Sore Risk
1
When the “Braden Scale” button is pressed on the
iBed screen, Figure 10 .22a will be displayed .
The Braden Scale is a 6 question survey the caregiver may use to evaluate the potential that a patient may develop pressure ulcers .
When the is pressed for any items 1-4, additional information will be provided for that item .
See example for item 2 (Very Limited) in Figure
10 .22b .
The survey results is a number (score) that is used to identify the care needed on the patient’s skin
(see Figure 10 .22c) . The lower the score the more attention the caregiver needs to provide to the patient’s skin .
When a Braden Scale is done to completion the last
Braden score is displayed in the Main screen with the date of the evaluation . The display of this score on the main screen can be enable or disabled .
Figure 10.22a
Figure 10.22b
1 Copyright . Barbara Braden and Nancy
Bergstrom, 1988 . Reprinted with permission .
Return To Table of Contents
1-56 2131-309-005 REV B
Figure 10.22c
www.stryker.com
Footboard Operation Guide
MAIN MENU: iBED (CONTINUED)
iBed, Sound Therapy™
When the “Sound Therapy™” button is pressed on the iBed screen, Figure 10 .23a will be displayed .
The operator may select a Sound Therapy™ from the Song(s) list and press play to activate . The volume of the sound selected can be adjusted by moving the volume slider up or down .
The Sound Therapy™ can be set to stop after a specified time which is calculated in minutes . Press the “Stop After” button to enable the stop time .
The playback mode allows the operator to play a single sound file or choose to play all files listed .
Pressing the Edit Play List button will display the
Play List screen as shown in Figure 10 .23b . The operator may add songs from the play list to the song list or remove songs from the song list .
If Sound Therapy™ is playing, a “Music Note” icon will be displayed at the bottom of the Main menu bar next to the date .
Note: The “Music Note” icon will continue to be displayed on all menu screens if Sound Therapy™ is playing .
Figure 10.23a
Figure 10.23b
Music Note Icon
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-57
ZOOM ® Drive Operations (Model 2141)
USING THE ZOOM
®
DRIVE (MODEL 2141)
1 . Disconnect the bed power cord from the wall socket .
2 . Push the Drive button (see Figure 24) located on either the head end siderail or on the footboard .
3 . The button will backlight and the LED will light up .
Figure 24
Note
• Handles might be hidden in the head section (see
Figure 25) .
• To take them out you have to unfold them and pull up to make sure they lock into place (See
Figure 26) .
4 . Use handles as follows:
• To move the bed forward, push on the handles while holding the yellow triggers .
• To move the bed backward, pull on the handles while holding the yellow triggers .
• To stop the movement, release both yellow triggers and stop pushing or pulling on the bed .
Note
• The speed of the bed’s movement will change according to the strength used to push or pull .
• If the bed battery charge level become too low, the motorized wheel will retract (raise) while driving, or won’t go down .
Figure 25
Figure 26
Return To Table of Contents
1-58 2131-309-005 REV B www.stryker.com
Optional Bed Accessories
The accessories listed below can be purchased and installed on the FL27 Intouch Critical Care Bed . Details on installation and service are included in the FL27 InTouch Critical Care Maintenance Manual .
Accessory
Bed Extender
Bed Extender Mattress (2920 PositionPRO only)
Bed Extender Mattress (2950 XPRT™ only)
I .V . Pole, Permanent - Left
I .V . Pole, Permanent - Right
I .V . Pole, Dual Head End Permanently Attached
Line Management System
Monitor Tray
Oxygen Bottle Holder, Upright
Oxygen Bottle Holder, Right Fit
Pendant with Motion Control Only
Pendant with Motion Control & Nurse Call
Pendant with Motion Control & Smart TV
Pendant with Motion Control, Nurse Call & Smart TV
Pendant Clip
Traction Sleeve 4” x 1/2”
Traction Sleeve 4” x 3/4”
Traction Sleeve 8” x 1/2”
Traction Sleeve 8” x 3/4”
Traction Sleeve 6-1/2” x 3/4”
Wall Saver Cable
X-Ray Cassette Holder
Part Number Page Reference
FA64234-XXX See page 1-60
DM64196
DM64197
See page 1-60
See page 1-60
FA64221
FA64238
See page 1-61
See page 1-61
FA64202
FA64210
FA64214
FA64187
FA64203
See page 1-63
See page 1-65
See page 1-66
See page 1-68
See page 1-69
FA64209
FA64194
FA64195
FA64193
FA64186
See page 1-70
See page 1-71
See page 1-72
See page 1-73
See page 1-74
FA64215
FA64216
FA64217
FA64218
FA64219
See page 1-75
FA64208
FA64205
See page 1-76
See page 1-77
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-59
Bed Extender - FA64234-XXX
The Optional Bed Extender was designed to lengthen the bed by six inches .
This accessory is designed for the following product(s):
InTouch 2 .5 (FL27)-2131
InTouch 2 .5 (FL27)-2141
INSTALLATION
Required Tools: None
Refer to the figures following the Installation section for visual help .
1 . Install the appropriate mattress cushion over the bed extender (see Figure A) .
2 . Remove the footboard .
3 . Insert bed extender legs on the bed end rods (see Figure B) .
4 . Replace the footboard over the bed extender rods (see Figure
B, section 2) .
WARNING
• Ensure connector ‘A’ is securely fixed (see Figure B) .
• Do not sit on the bed extender . This may cause the bed to flip .
Figure A
Section 2
Connector A
Return To Table of Contents
1-60
Figure B
2131-309-005 REV B
Section 1 www.stryker.com
Permanent I.V. Pole - FA64221/FA64238
The Optional Single 2-Stage I .V . Pole is permanently attached and equipped with a telescopic pole . The telescopic pole is an extension which provides a second height position . The Permanent I .V . Pole can be folded and stored .
INSTALLATION
Required Tools:
• 3/8” Drive Ratchet
• 1/2” Socket
1 . Using bolt (D) and washer (C), secure the I .V . pole (A) into the appropriate opening to the left of the head of the bed .
2 . Using bolt (D) and washer (C), secure the I .V . pole support (B) in the appropriate opening to the right of the head of the bed .
Note: The bolt used is coated with “Scotch Grip”. This type of bolt must be replaced by a new identical bolt if it is removed following a first installation.
A
B
C
D
FA64221 - Left
-Bwww.stryker.com
-A-
-C-
-D-
FA64238 - Right
2131-309-005 REV B
Return To Table of Contents
1-61
Permanent I.V. Pole (Continued) - FA64221/FA64238
OPERATION
1 . Lift and pivot the pole from the storage position and push down until it is locked into the receptacle .
2 . To raise the height of the pole, pull up on the telescoping portion (A) until it locks into place at its fully raised position . There are no intermediary positions for this section .
3 . Rotate the I .V . pole hangers (B) to desired position and hang the I .V . bag .
4 . To store the I .V . pole, turn the latch (C) clockwise until section (A) lowers . Lift up on the pole to remove it from its receptacle, fold it and store it on the bracket .
CAUTION
The weight of the I .V . bags should not exceed 22 lbs (10 kg) .
Return To Table of Contents
1-62 2131-309-005 REV B www.stryker.com
Dual Head End Permanent I.V. Pole - FA64202
The Optional Dual Head End I .V . Pole is permanently attached and equipped with a telescopic pole . The telescopic pole is an extension which provides a second height position . The Permanent I .V . Pole can be folded and stored .
INSTALLATION
Required Tools:
• 3/8” Drive Ratchet
• 1/2” Socket
1 . Using bolt (D) and washer (C), secure the I .V . pole (A) into the appropriate opening to the left of the head of the bed .
2 . Using bolt (D) and washer (C), secure the I .V . pole support (B) in the appropriate opening to the right of the head of the bed .
Note: The bolt used is coated with “Scotch Grip”. This type of bolt must be replaced by a new identical bolt if it is removed following a first installation.
A
C
D
B www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-63
Dual Head End Permanent I.V. Pole (Continued) - FA64202
OPERATION
1 . Lift and pivot the pole from the storage position and push down until it is locked into the receptacle .
2 . To raise the height of the pole, pull up on the telescoping portion (A) until it locks into place at its fully raised position . There are no intermediary positions for this section .
3 . Rotate the I .V . pole hangers (B) to desired position and hang the I .V . bag .
4 . To store the I .V . pole, turn the latch (C) clockwise until section (A) lowers . Lift up on the pole to remove it from its receptacle, fold it and store it on the bracket .
CAUTION
The weight of the I .V . bags should not exceed 22 lbs (10 kg) .
Return To Table of Contents
1-64 2131-309-005 REV B www.stryker.com
Line Management System (Accessory Clip) - FA64210
The Optional Line Management System (Accessory Clip) was designed to help manage the tubes around the bed .
INSTALLATION
Required Tools: None
1 . Open clip (A) and install at a strategic and secure location on the bed .
2 . Raise clip prong (B) and insert tubes of ¾” diameter or less .
3 . Close opening and make sure that the tubes are not pinched .
WARNING
• Make sure the clip is not located where it can interfere with a mechanism or activate one .
• If accessory is placed in a liquid solution it could deteriorate the mechanism .
CAUTION
• Do not use with tubing of a diameter greater than 0 .75” (19 mm) .
• Sterilize accessory after each use (see procedure in Maintenance Manual)
• Make sure that the clip is stable when installing .
A
B www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-65
Monitor Tray - FA64214
The Optional Monitor Tray is designed to hold a monitor at the foot end of the bed .
WARNING
The Weigh system as well as the Bed Exit system must be adjusted if this accessory is added when either system is in function . Refer to the “Add/Remove Equipment” from the bed procedure on page 33 .
INSTALLATION
Required Tools: None
Screw Knob Bracket
Procedure
1 . Install the bracket (A) at the left side of the foot end . (see
Figures B, C, D, and E) .
2 . Install the transport tray (see Figure F) into the bracket just secured in step 1 and into the the IV socket (not shown) .
Note: Ensure that the tray support is correctly anchored in the hole of the bracket installed in step 1 (see Figure F).
Figure A
Figure B - foot end Figure C - bracket mounted (back) Figure D - bracket mounted (front)
Figure E - secure bracket with screw knob
Return To Table of Contents
1-66
Figure F - install tray
2131-309-005 REV B www.stryker.com
Monitor Tray (Continued) - FA64214
OPERATION
1 . Pull the lock down (see Figure G) and turn the tubular support 90° (see Figure H) to get it perpendicular to the board (see Figure I) then release the lock .
2 . Hold the bottom of the plastic tray and flip it up and over so it rests on the tubular support . Apply a pressure on the tray to stabilize it on the tubular suppport (see Figure J) .
3 . Secure the monitor to the tray using the strap (see Figure K, L) .
Note: The litter should be flat before installing the transport tray.
Figure G - pull the lock Figure H - turn the tubular support Figure I - pendicular to board
Figure J - stabilize tray Figure K - secure with straps Figure L - secure with straps
WARNING
• The maximum load capacity of the tray is 40 lbs (18kg)
• The maximum load capacity of the tubular support is 150 lbs (68 kg) www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-67
Upright Oxygen Bottle Holder - FA64187
The Optional Upright Oxygen Bottle Holder was designed to support the oxygen bottle .
INSTALLATION
Required Tools: None
1 . Insert the Support Bar in the middle hole on the right or left side of the head end of the bed .
2 . Secure the Support Bar to the head end of the bed by inserting the security chain through the Support Bar hole located at the end of the Support Bar .
Support Bar
Support Bar Hole
Security
Chain
Return To Table of Contents
1-68 2131-309-005 REV B www.stryker.com
Right Fit Oxygen Bottle Holder - FA64203
The Optional Right Fit Oxygen Bottle Holder is designed to support the oxygen bottle on the headboard .
INSTALLATION
Required Tools: None
1 . Install bottle holder (A) on headboard (B) .
2 . Screw the holder handle (C) until the holder is installed properly on the headboard .
3 . Route the number 3 straps (D and E) through the holes in the headboard and attach them to their fasteners .
4 . Insert the oxygen bottle in the holder .
5 . Install the cover (F) on the holder and tie the number 1 straps (G) together .
Note: The cover is reversible. It can be placed so that the opening is on the right or on the left of the bed to ease the use.
G
E
F
Insert the oxygen bottle here .
A
D
C
B www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-69
Pendant with Motion Control Only - FA64209
The Optional Pendant with Motion Control was designed to give the patient control over the bed’s motion .
Motion Control
WARNING
The Weigh system as well as the Bed Exit system must be adjusted if this accessory is added when either system is in function . Refer to the “Add/Remove Equipment” from the bed procedure on page 33 .
INSTALLATION
Required Tools: None
1 . Plug the cable connector into connector (A) located on the right of the upper frame .
2 . Install the removable pendant on the right foot siderail .
Insert the removable pendant in in this space. Slide it until it clips.
A
Return To Table of Contents
1-70
Note: Illustration may differ according to bed model.
2131-309-005 REV B www.stryker.com
Pendant with Motion Control/Nurse Call - FA64194
The Optional Pendant with Motion Control and Nurse Call was designed to give the patient control over the bed’s motion and the ability to contact the nurses from the bed when needed .
The bed must be configured and synchronized with the hospital system in order for the Nurse Call function to be available .
Nurse Call
Motion Control
WARNING
The Weigh system as well as the Bed Exit system must be adjusted if this accessory is added when either system is in function . Refer to the “Add/Remove Equipment” from the bed procedure on page 33 .
INSTALLATION
Required Tools: None
1 . Plug the cable connector into connector (A) located on the right of the upper frame .
2 . Install the removable pendant on the right foot siderail .
Insert the removable pendant in in this space. Slide it until it clips.
A www.stryker.com
Note: Illustration may differ according to bed model.
2131-309-005 REV B
Return To Table of Contents
1-71
Pendant with Motion Control/Smart TV - FA64195
The Optional Pendant with Motion Control and Smart TV was designed to give the patient control over the bed’s motions, TV, radio, channels, volume and lights from the bed when needed .
The bed must be configured and synchronized with the hospital system in order for some functions to operate properly such as TV control, radio, channels, volume and lights .
Motion Control
Smart TV
WARNING
The Weigh system as well as the Bed Exit system must be adjusted if this accessory is added when either system is in function . Refer to the “Add/Remove Equipment” from the bed procedure on page 33 .
INSTALLATION
Required Tools: None
1 . Plug the cable connector into connector (A) located on the right of the upper frame .
2 . Install the removable pendant on the right foot siderail .
Insert the removable pendant in in this space. Slide it until it clips.
A
Return To Table of Contents
1-72
Note: Illustration may differ according to bed model.
2131-309-005 REV B www.stryker.com
Pendant with Motion Control/NC/Smart TV - FA64193
The Optional Pendant with Motion Control, Nurse Call and Smart TV was designed to give the patient control over the bed’s motions, TV, radio, channels, volume and lights as well as being able to contact a nurse from the bed when needed .
The bed must be configured and synchronized with the hospital system in order for some functions to operate properly such as Nurse Call, TV control, radio, channels, volume and lights .
Nurse Call
Smart TV
WARNING
The Weigh system as well as the Bed Exit system must be adjusted if this accessory is added when either system is in function . Refer to the “Add/Remove Equipment” from the bed procedure on page 33 .
INSTALLATION
Motion Control
Required Tools: None
1 . Plug the cable connector into connector (A) located on the right of the upper frame .
2 . Install the removable pendant on the right foot siderail .
Insert the removable pendant in in this space. Slide it until it clips.
A www.stryker.com
Note: Illustration may differ according to bed model.
2131-309-005 REV B
Return To Table of Contents
1-73
Pendant Clip - FA64186
The Optional Pendant Clip was designed to support the removable pendant (FA64193, FA64194, FA64195, FA64209) .
INSTALLATION
Required Tools: None
1 . Raise the clip prong (A) and install the clip on the bed at a safe and useful location .
A
Return To Table of Contents
1-74 2131-309-005 REV B www.stryker.com
Traction Sleeve - FA64215 through FA64219
The Optional Traction Sleeve enables the installation of traction equipment . There are five different size configurations available as shown below . The installation procedures below apply to all configurations listed .
NOTE: The foot end section of the mattress is stopped when a traction socket or I .V . pole is inserted .
• Traction Sleeve Set - 4” x 1/2”
• Traction Sleeve 4” x 3/4”
• Traction Sleeve 8” x 1/2”
• Traction Sleeve 8” x 3/4”
• Traction Sleeve 6-1/2” x 3/4”
INSTALLATION
FA64215
FA64216
FA64217
FA64218
FA64219
Required Tool:
• 7/16” Combination Wrench
Procedure
1 . Install the four traction sleeves (A) in the corresponding holes located at the four corners of the bed .
2 . Secure them using the washers (B) and bolts (C) provided .
Note: The bolt used at step 2 is coated with “Scotch Grip”. This type of bolt must be replaced by a new identical bolt if it is removed following a first installation.
A
A
B www.stryker.com
C
B
2131-309-005 REV B
C
B
B
Return To Table of Contents
1-75
Wall Saver - FA64208
The Optional Wall Saver is designed to avoid damages while moving the bed if the bed is not unplugged . This wall saver will easily unconnect itself .
INSTALLATION
Required Tools: None
Procedure
1 . Insert the connector in the outlet on the bed, as indicated in the figure below .
2 . Screw the connector’s screws at the bed location .
3 . Join both cables .
4 . Insert the connector in the other one on the cable from the hospital .
5 . Screw the connector’s screws to the hospital connector .
Return To Table of Contents
1-76 2131-309-005 REV B www.stryker.com
X-Ray Cassette Holder - FA64205
The Optional X-Ray Cassette Holder was designed to hold the necessary material in order to perform X-Rays with the bed .
INSTALLATION
Required Tools:
• #2 Phillips Screwdriver
1 . Raise the fowler to the full upright position .
2 . Apply the brakes if they are not already engaged .
3 . Unplug the bed and turn off the battery disconnect switch .
4 . Using a Phillips screwdriver, install the bracket assembly (A) to the the top of the fowler frame with two #10 x ½” tapping screws (F) .
5 . Using a Phillips screwdriver, install the two Cassette Holder Pivot brackets (B) into the lower section of the fowler with four #10 x ½” tapping screws (G) .
6 . Using the Phillips screwdriver, install the X-Ray Cassette Holder Plate (C) into the Cassette Holder Pivot brackets
(B) with two #10 x 3/4” tapping screws (E) and two spacers (D)
Note: Make sure to close the X-ray cassette holder support after using it to avoid damaging it when there is a movement of the fowler section or of the list system.
G
B
D E
D
C
B
G
F
A www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-77
Cleaning
BED CLEANING
Hand wash all surfaces of the bed with warm water and mild detergent . Wipe the bed clean and DRY THOROUGHLY to avoid build up of cleaning solution . Do not steam clean or hose off the bed . Do not immerse any part of the bed .
Some of the internal parts of the bed are electric and may be damaged by exposure to water .
Suggested cleaners for bed surfaces:
• Quaternary Cleaners (active ingredient - ammonium chloride) .
• Phenolic Cleaners (active ingredient - o-phenylphenol) .
• Chlorinated Bleach Solution (5 .25% - less than 1 part bleach to 100 parts water) .
Avoid over saturation and ensure the product does not stay wet longer than the chemical manufacturer’s guidelines for proper disinfecting .
CAUTION
SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF USED
IMPROPERLY . If the products described above are used to clean Stryker patient care equipment, measures must be taken to ensure the beds are wiped with a damp cloth soaked in clean water and thoroughly dried following cleaning .
Failure to properly rinse and dry the beds will leave a corrosive residue on the surface of the bed, possibly causing premature corrosion of critical components . Failure to follow the above directions when using these types of cleaners may void this product’s warranty .
MATTRESS CARE
Inspect the mattress after each use . Discontinue use if any cracks or rips are found in the mattress cover as these openings may allow fluid to enter the mattress .
Return To Table of Contents
1-78 2131-309-005 REV B www.stryker.com
Preventative Maintenance
Beds require an effective maintenance program, we recommend checking these items annually . Use this sheet for your records . Keep on file .
CHECKLIST
_____ All fasteners secure (reference all assembly prints) .
_____ Engage brake pedal and push on the bed to ensure all casters lock securely .
_____ “Brake Not Set” LED on the footboard and head end siderails when brakes are not engaged .
_____ Locking steer caster engages and disengages properly (2131 Model only) .
_____ Siderails move, latch and stow properly .
_____ CPR release working properly .
_____ I .V . pole working properly (if on bed) .
_____ No cracks or splits in head or footboards .
_____ No rips or cracks in mattress cover .
_____ All functions on head end siderails working properly (including LED’s) .
_____ All functions on footboard working properly (including LED’s) .
_____ Scale and bed exit system working properly .
_____ Night light working properly .
_____ Power cord not frayed .
_____ No cables worn or pinched .
_____ All electrical connections tight .
_____ All grounds secure to the frame .
_____ Ground impedance not more than 100 milliohms .
_____ Current leakage not more than 300 microamps .
_____ Engage drive wheel and ensure it is operating properly (ZOOM® option - 2141 model only) .
_____ Motion release switches working properly (ZOOM® option - 2141 model only) .
_____ Confirm head end ZOOM® handle functionality (2141 model only) .
_____ Confirm battery powered functionality .
_____ Ensure ground chains are clean, intact, and have at least two links touching the floor .
_____ Check fowler angle for accuracy 0 0 - 70 0 .
_____ Check gatch angle for accuracy 0 0 - 15 0 .
_____ Check foot section angle for accuracy 0 0 - 50 0 .
_____ Siderail switches working properly (iBED Awareness option) .
_____ iBED Awareness Light Bar LED’s on footboard and siderails working properly (iBED Awareness option) .
_____ Inspect footboard control labeling for signs of degradation .
_____ Ensure calibration of the touch screen is accurate .
_____ Ensure calibration of the bed is accurate .
Bed Serial Number: ________________________ ________________________
________________________ ________________________
________________________ ________________________
________________________ ________________________
Completed by: ___________________________________________ Date: _________________ www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-79
Recycling Passports
Assembly Part number: 27-2287/27-2288 (Reference Only)
A
B
(27-2287 shown)
Item
A
B
Recycling/Material Code
(QDF27-1097) Brake Control Board
(QDF27-1099) Siderail
Nurse Control Board
Important Information Qty
2
2
Return To Table of Contents
1-80 2131-309-005 REV B www.stryker.com
Recycling Passports
Assembly Part number: OL270254 (Reference Only)
A
B
Item
A
B
Recycling/Material Code
(QDF27-1429) Nurse Call Board
(QDF75-0410) Can-Conn Board www.stryker.com
Important Information
2131-309-005 REV B
Qty
1
1
Return To Table of Contents
1-81
Recycling Passports
Assembly Part number: OL270032 (Reference Only)
A
Item
A
Recycling/Material Code
(QDF2060) Smart TV Board
Important Information Qty
1
Return To Table of Contents
1-82 2131-309-005 REV B www.stryker.com
Recycling Passports
Assembly Part number: OL270060 (Reference Only)
A
A
Item
A
Recycling/Material Code
(QDF27-1562) iBed Electronic Board Lens
Important Information Qty
4 www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-83
Recycling Passports
Assembly Part number: OL270263-XXX (Reference Only)
A
Item
A
Recycling/Material Code
(QDF27-1099) Siderail Nurse Control Board
Important Information Qty
3
Return To Table of Contents
1-84 2131-309-005 REV B www.stryker.com
Recycling Passports
Assembly Part number: 27-2547-XXX (Reference Only)
A
B
Item
A
B www.stryker.com
Recycling/Material Code
QDF75-0230 (ZOOM Control Board)
QDF27-1372 (ZOOM Load Cell)
Important Information
2131-309-005 REV B
Qty
1
1
Return To Table of Contents
1-85
Recycling Passports
Assembly Part number: OL270293-XXX (Reference Only)
A
Item
A
Recycling/Material Code
(QDF27-1562) iBed Electronic Board Lens
Important Information Qty
4
Return To Table of Contents
1-86 2131-309-005 REV B www.stryker.com
Recycling Passports
Assembly Part number: OL270294-XXX (Reference Only)
A
A
B
Item
A
B
Recycling/Material Code
(QDF75-0010) Menu Board
(QDF27-1097) Brake Control Board
Important Information Qty
1
1 www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-87
Recycling Passports
Assembly Part number: L27-062 (Reference Only)
A
Item
A
Recycling/Material Code
(27-2477) Fowler Angle Sensor
Important Information Qty
1
Return To Table of Contents
1-88 2131-309-005 REV B www.stryker.com
Recycling Passports
Assembly Part number: L27-062 (Reference Only)
A
Item
A
Recycling/Material Code
(27-2477) Gatch Angle Sensor
Important Information Qty
1 www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-89
Recycling Passports
Assembly Part number: 27-2688 (Reference Only)
A
A
Item
A
Recycling/Material Code
(27-2477) Foot Angle Sensor
Important Information
Return To Table of Contents
1-90 2131-309-005 REV B
Qty
1 www.stryker.com
Recycling Passports
Assembly Part number: L27-043 (Reference Only)
Item
A
Recycling/Material Code
(27-2477) Base Angle Sensor
Important Information
A
Qty
1 www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-91
Recycling Passports
Assembly Part number: L27-046 (Reference Only)
A
Item
A
Recycling/Material Code
(27-2477) Trend Angle Sensor
Important Information Qty
1
Return To Table of Contents
1-92 2131-309-005 REV B www.stryker.com
Recycling Passports
Assembly Part Number: 27-2688 (Reference Only)
B A
Item
A
B
Recycling/Material Code
(QDF75-0450) DC Power Control Board
(QDF9188) 12V 18Ah Battery
Important Information Qty
1
1 www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-93
Recycling Passports
Assembly Part number: 27-2546 (Reference Only)
B
A
Item
A
B
Recycling/Material Code
(QDF27-1430) ZOOM Board
(QDF75-0240) ZOOM Interface Board
Return To Table of Contents
1-94
Important Information
2131-309-005 REV B
Qty
1
1 www.stryker.com
Recycling Passports
Assembly Part number: L27-055 (Reference Only)
A
Item
A
Recycling/Material Code
(QDF27-1372) Litter Load Cell
Important Information www.stryker.com
2131-309-005 REV B
Qty
4
Return To Table of Contents
1-95
Warranty
LIMITED WARRANTY
Stryker Medical Division, a division of Stryker Corporation, warrants to the original purchaser the InTouch® Critical
Care Bed, Model FL27 (2131/2141) be free from defects in material and workmanship for a period of one (1) year after date of delivery . Stryker’s obligation under this warranty is expressly limited to supplying replacement parts and labor for, or replacing, at its option, any product which is, in the sole discretion of Stryker, found to be defective . If requested by Stryker, products or parts for which a warranty claim is made shall be returned prepaid to the factory . Any improper use or any alteration or repair by others in such manner as in Stryker’s judgment affects the product materially and adversely shall void this warranty . Any repair of Stryker products using parts not provided or authorized by Stryker shall void this warranty . No employee or representative of Stryker is authorized to change this warranty in any way .
Stryker Medical Bed products are designed for a 10 year expected service life under normal use, conditions, and with appropriate periodic maintenance as described in the maintenance manual for each device . Stryker warrants to the original purchaser that the welds on its Bed products will be free from structural defects for the expected 10 year life of the Bed product as long as the original purchaser owns the product .
This statement constitutes Stryker’s entire warranty with respect to the aforesaid equipment . Stryker makes no other warranty or representation, either expressed or implied, except as set forth herein. There is no warranty of merchantability and there are no warranties of fitness for any particular purpose. In no event shall Stryker be liable here under for incidental or consequential damages arising from or in any manner related to sales or use of any such equipment.
TO OBTAIN PARTS AND SERVICE
Stryker products are supported by a nationwide network of dedicated Stryker Field Service Representatives . These representatives are factory trained, available locally, and carry a substantial spare parts inventory to minimize repair time . Simply call your local representative, or call Stryker Customer Service USA at 1-800-327−0770, option 2, Canada
1-888-233-6888 .
SERVICE CONTRACT COVERAGE
Stryker has developed a comprehensive program of service contract options designed to keep your equipment operating at peak performance at the same time it eliminates unexpected costs . We recommend that these programs be activated before the expiration of the new product warranty to eliminate the potential of additional equipment upgrade charges .
A Service Contract helps to:
• Ensure equipment reliability
• Stabilize maintenance budgets
• Diminish downtime
• Establish documentation for JCAHO
• Increase product life
• Enhance trade-in value
• Address risk management and safety
Return To Table of Contents
1-96 2131-309-005 REV B www.stryker.com
Warranty
SERVICE CONTRACT PROGRAMS
Stryker offers the following service contract programs:
Service Agreement Options *
Annually scheduled preventative maintenance
All parts
All labor and travel
Unlimited emergency service calls
Priority one contact: two hour phone response
Most repairs completed within 3 days
JCAHO documentation
On-site record of PM & emergency service
Factory-trained Stryker service technician
Stryker authorized parts used
Service during regular business hours (8−5)
Premium
X
X
X
X
X
X
X
X
X
X
X
Complete
X
X
X
X
X
X
X
X
X
Standard
* Does not include maintenance due to abuse or for any disposable items . Stryker reserves the right to change options without notice .
X
X
X
X
X
X
Stryker Medical also offers personalized service contracts .
Pricing is determined by age, location, model and condition of product .
For more information on our service contracts, please call your local representative.
RETURN AUTHORIZATION
Merchandise cannot be returned without approval from the Stryker Customer Service Department . An authorization number will be provided which must be printed on the returned merchandise . Stryker reserves the right to charge shipping and restocking fees on returned items . Special, modified, or discontinued items not subject to return.
DAMAGED MERCHANDISE
ICC Regulations require that claims for damaged merchandise must be made with the carrier within fifteen (15) days of receipt of merchandise . Do not accept damaged shipments unless such damage is noted on the delivery receipt at the time of receipt. Upon prompt notification, Stryker will file a freight claim with the appropriate carrier for damages incurred . Claim will be limited in amount to the actual replacement cost . In the event that this information is not received by Stryker within the fifteen (15) day period following the delivery of the merchandise, or the damage was not noted on the delivery receipt at the time of receipt, the customer will be responsible for payment of the original invoice in full . Claims for any short shipment must be made within thirty (30) days of invoice .
INTERNATIONAL WARRANTY CLAUSE
This warranty reflects U .S . domestic policy . Warranty outside the U .S . may vary by country . Please contact your local
Stryker Medical representative for additional information .
www.stryker.com
2131-309-005 REV B
Return To Table of Contents
1-97
Download
Advertisement
Key features
Patient positioning for procedures, therapy, and recovery
Patient transport between bays and procedural rooms
Patient security
Patient weight measurement and display
Nurse call functionality
Siderail communications
Bed exit detection
Therapy options (e.g., music therapy)
Frequently asked questions
The safe working load for the InTouch FL27 bed is 550 lbs (250 kg).
The overall length of the InTouch FL27 bed is 90 inches (228.6 cm).
The maximum current consumption of the InTouch FL27 bed is 4.8 Amps without the optional auxiliary outlet and 9.8 Amps with the optional auxiliary outlet.
The recommended mattress sizes vary depending on whether you are using a bed extender. Please refer to the specifications section of the manual for detailed information.