Agilent U1401A Handheld Multi-function Calibrator/Meter


Add to my manuals
8 Pages

advertisement

Agilent U1401A Handheld Multi-function Calibrator/Meter | Manualzz

Datasheet

Agilent U1401A Handheld Multi-function

Calibrator/Meter

Data Sheet

Calibrate while you measure with just one tool

Agilent U1401A handheld multi-function calibrator/meter has all you need for quick validation, servicing or troubleshooting of process control devices on the go. Travel and test easily with one rugged, featurepacked tool.

Features

Dual display with bright LCD backlight

Simultaneous source and measure

Bipolar voltage and current,

square-wave, auto scan and

ramp outputs

Full-span DMM capability, including

temperature and frequency

measurements

Hold and Min/Max/Average recordings

Data logging to PC with optional

IR-to-USB cable

Built-in charging capability

The 2-in-1 that helps you travel light

More often than not, the calibration of process control parts requires simultaneous measurements with a DMM. With the

U1401A, you can carry two tools in one— and calibrate while you measure. Slip the

U1401A in its sturdy carrying case and you’re ready to go.

Full-featured DMM functions

The U1401A is packed with a full span of

DMM measurement functions, including

AC+DC voltage and current, resistance, temperature, frequency, diode and continuity tests. It also equips you with recording functions such as Hold, Min/Max/Average and data logging to PC.

Rugged and tested to stringent standards

The U1401A comes with a robust protective holster and tested to stringent industrial standards. Each U1401A is also sealed with a three-year warranty and the assurance that you can perform your calibration tasks with confidence.

Amplicon.com

IT and Instrumentation for industry

Sales: +44 (0) 1273 570 220 Website: www.amplicon.com Email: [email protected]

Datasheet

Take a closer look

Robust protective holster

Hold and Min/Max/Avg recordings

Built-in batterycharging capability

Bipolar voltage and current, square-wave, auto scan and ramp

Dual display with backlight

4-20 mA and 0-20 mA percentage scale readout

Easy function access through intuitive buttons and rotary switch

Simultaneous source and measure

Voltage, current, resistance, temperature, frequency, pulse width, duty cycle, diode and continuity

Amplicon.com

IT and Instrumentation for industry

Sales: +44 (0) 1273 570 220 Website: www.amplicon.com Email: [email protected]

Datasheet

Input Specifications

The accuracy is given as ± (% of reading + counts of least significant digit (LSD)) at 23 °C ±

5 °C, with relative humidity less than 80% R.H., and after a warm-up period of at least five minutes.

Without warm-up, an additional five counts of LSD need to be considered.

Voltage specifications

Function

DC voltage [1]

Range

50 mV

500 mV

5 V

50 V

250 V

Resolution

1 µV

10 µV

0.1 mV

1 mV

10 mV

Accuracy

0.05% + 50 [2]

0.03% + 5

AC voltage [3]

(True-rms: From 5% to

100% of range)

50 mV 1 µV

45 Hz to 5 kHz: 0.7% + 40

5 kHz to 20 kHz: 1.5% + 40

500 mV

5 V

50 V

250 V

10 µV

0.1 mV

1 mV

10 mV

45 Hz to 5 kHz: 0.7% + 20

5 kHz to 20 kHz: 1.5% + 20

AC+DC voltage [3]

(True-rms: From 5% to

100% of range)

50 mV 1 µV

45 Hz to 5 kHz: 0.8% + 70

5 kHz to 20 kHz: 1.6% + 70

500 mV

5 V

50 V

250 V

10 µV

0.1 mV

1 mV

10 mV

45 Hz to 5 kHz: 0.8% + 25

5 kHz to 20 kHz: 1.6% + 25

Overload protection

250 Vrms

[1] Input impedance: 10 MW (nominal) for the range of 5 V and above, and 1 GW (nominal) for the 50/500 mV range.

[2] Accuracy can be improved to 0.05% + 5. Always use the Relative function to offset thermal effects before measuring the signal.

[3] Input impedance: 1.1 MW in parallel with <100 pF (nominal) for the range of 5 V and above, and 1 GW (nominal) for the 50/500 mV range. Crest factor ~ 3.

Amplicon.com

IT and Instrumentation for industry

Sales: +44 (0) 1273 570 220 Website: www.amplicon.com Email: [email protected]

Datasheet

Current specifications

Function

DC current

AC current [2]

(True-rms: From 5% to 100% of range)

AC+DC current [2]

(True-rms: From 5% to 100% of range)

Range

50 mA

500 mA

50 mA

500 mA

50 mA

500 mA

Resolution

1 µA

10 µA

1 µA

10 µA

1 µA

10 µA

Accuracy

0.03% + 5 [1]

45 Hz to 5 kHz:

0.6% + 20

45 Hz to 5 kHz:

0.7% + 25

Burden voltage/shunt Overload protection

0.06 V (1 W)

0.6 V (1 W)

0.06 V (1 W)

0.6 V (1 W) 250 V, 630 mA

Quick acting fuse

0.06 V (1 W)

0.6 V (1 W)

[1] Always use the Relative function to offset thermal effects before measuring the signal. If this function is not used, accuracy

could go down to 0.03% + 25. Thermal effects may be present due to:

• Constant current, constant voltage, or square wave output.

• Wrong operation. For example, resistance, diode, or mV measurement function is used to measure high voltage signals exceeding 250 V.

• After battery charging has completed.

• After measuring current greater than 50 mA.

[2] Crest factor ~ 3

Amplicon.com

IT and Instrumentation for industry

Sales: +44 (0) 1273 570 220 Website: www.amplicon.com Email: [email protected]

Datasheet

Temperature specifications

Thermocouple type

K

Range

–40 °C to 1372 °C

–40 °F to 2502 °F

Resolution

0.1 °C

0.1 °F

Accuracy [1]

0.3% + 3 °C

0.3% + 6 °F

Overload protection

250 Vrms

Resistance specifications

The following resistance specifications are valid if the maximum open voltage is less than +4.8 V.

Range

500 W

5 kW

50 kW

500 kW

5 MW

50 MW

Resolution

0.01 W

0.1 W

1 W

10 W

0.1 kW

1 kW

Accuracy

0.15% + 8 [2]

0.15% + 5 [2]

1% + 8 [3]

Minimum input current

0.45 mA

0.45 mA

45 µA

4.5 µA

450 nA

45 nA

Diode and continuity specifications

For diode test, the overload protection is 250 Vrms and the instrument will beep when the reading is below 50 mV (approx). For continuity test, the instrument will beep when the resistance is less than 10.00 W.

Resolution

0.1 mV

Accuracy

0.05% + 5

Test current

Approximately 0.45 mA

Open voltage

< +4.8 VDC

1 ms peak hold specifications

Signal width

Single event >1 ms

Accuracy for DC mV/voltage/current

2% + 400 for all ranges

Overload protection

250 V rms

[1] Accuracy is specified for meter operation only, excludes thermocouple probe tolerance and with the instrument placed in the operating area for at least one hour.

[2] Accuracy is specified after applying the Relative function to offset any test lead resistance and thermal effect.

[3] Accuracy is specified for <60% R.H.

Amplicon.com

IT and Instrumentation for industry

Sales: +44 (0) 1273 570 220 Website: www.amplicon.com Email: [email protected]

Datasheet

Frequency specifications

Range

100 Hz

1 kHz

10 kHz

100 kHz

200 kHz

Resolution

0.001 Hz

0.01 Hz

0.1 Hz

1 Hz

10 kHz

Accuracy

0.02% + 3

Minimum input frequency

1 Hz

Frequency sensitivity and trigger level for voltage measurement

Input range Minimum sensitivity

(rms sine wave)

50 mV

500 mV

5 V

50 V

250 V

1 Hz to 100 kHz

15 mV

35 mV

0.3 V

3V

30 V

>100 kHz

25 mV

50 mV

0.5 V

5 V

–

<20 kHz

20 mV

60 mV

0.6 V

6 V

60 V

Overload protection

250 Vrms

Trigger level for DC coupling

20 kHz to 200 kHz

30 mV

80 mV

0.8 V

8 V

–

Frequency sensitivity for current measurement

Input range

50 mA

500 mA

Minimum sensitivity (rms sine wave)

30 Hz to 20 kHz

2.5 mA

25 mA

Duty cycle and pulse width

Function

Duty cycle

Pulse width [2]

Mode

DC coupling

AC coupling

–

Range

0.1% to 99.9%

5% to 95%

0.01 ms to 1999.9 ms

[1] Accuracy is based on a 5-V square-wave input to the 5 VDC range.

[2] Pulse width must be greater than 10 µs and its range is determined by the frequency of the signal.

Accuracy at full scale [1]

0.3% per kHz + 0.3%

0.2% + 3

Amplicon.com

IT and Instrumentation for industry

Sales: +44 (0) 1273 570 220 Website: www.amplicon.com Email: [email protected]

Datasheet

Output Specifications

Accuracy is given as ± (% of output + counts of least significant digit (LSD)) at 23 °C ± 5 °C, with relative humidity less than 80% R.H., and after a warm-up period of at least five minutes.

The maximum input voltage protection is 30 VDC.

Constant voltage and current outputs

Function

Constant voltage (CV)

Constant current (CC)

Range

± 1.500 V

± 15.000 V

± 25.000 mA

Resolution

0.1 mV

1 mV

1 µA

Accuracy

0.03% + 3

0.03% + 5

Minimum output

25 mA or above [1]

12 V or above [2][3]

Square wave output

Output

Frequency (Hz)

Duty Cycle (%) [4]

Pulse Width (ms)

Amplitude (V)

[4]

Range

0.5, 1, 2, 5, 10, 15, 20, 25, 30, 40, 50, 60, 75,

80, 100, 120, 150, 200, 240, 300, 400, 480,

600, 800, 1200, 1600, 2400, 4800

0.39% to 99.60%

1/Frequency

5 V, 12 V

±5 V, ±12 V

Resolution

0.01

0.390625%

Range/256

0.1 V

Accuracy

0.005% + 1

0.01% + 0.2% [5]

0.01% + 0.3 ms

2% + 0.2 V

2% + 0.4 V

[1] Loading coefficient: 0.012 mV/mA for 1.5 V output.

[2] Loading coefficient: 1 µA/ V. The minimum output voltage is based on 20 mA into a 600 W load.

[3] If the current loop has a 24-V power, a minimum output voltage of 24 V is achievable with a 20 mA current into a 1200-W load (using the yellow test lead).

[4] The positive or negative pulse width must be greater than 50 µs to enable adjustment of duty cycle or pulse width under different frequencies. Otherwise, the accuracy and range will

be different from the specifications defined.

[5] For signal frequencies greater than 1 kHz, add an addition of 0.1% per kHz.

Amplicon.com

IT and Instrumentation for industry

Sales: +44 (0) 1273 570 220 Website: www.amplicon.com Email: [email protected]

Datasheet

General Specifications

Display

Power supply

Power consumption

Battery life

Charging time

Measurement rate

Common Mode Rejection Ratio (CMRR)

Normal Mode Rejection Ratio (NMRR)

Operating environment

Storage environment

Altitude

Safety compliance

EMC compliance

Temperature coefficient

Dimensions (H x W x D)

Weight

Calibration

Warranty

Both primary and secondary displays are 5-digit on the liquid crystal display (LCD) with a maximum resolution of 51,000 counts and automatic polarity indication. Backlight available.

9.6 V Ni-MH rechargeable batteries: 1.2 V x 8 pieces. No cadmium, lead or mercury.

External switching adapter: AC 100 V to 240 V, 50/60 Hz input and DC 24 V/2.5 A output

Battery charging: 9.3 VA typical

Sourcing of constant current at 25 mA, maximum load: 5.5 VA typical on 24 V DC adapter,

2.4 VA typical on 9.6 V batteries

Meter only: 1.8 VA typical on 24 V DC adapter, 0.6 VA typical on 9.6 V batteries

Assuming fully-charged Ni-MH batteries:

Meter only: 20 hours (approx.)

Source/Meter: 4 hours (approx.)

will appear when voltage drops below 9 V (approx.)

3 hours (approx.) in 10 °C to 30 °C environment

NOTE: Prolonged charging is required if battery is fully discharged.

3 readings/second, except for:

AC+DC: 1 reading/second

Frequency and duty cycle (> 1 Hz): 1 reading/second

Pulse width (> 1 Hz): 0.25 to 1 reading/second

> 90 dB at DC, 50/60 Hz ± 0.1% (1 kW unbalanced)

> 60 dB at DC, 50/60 Hz ± 0.1%

0 °C to 40 °C; up to 80% relative humidity (R.H.) for temperatures up to 31 °C, decreasing linearly to 50% R.H. at 40 °C

–20 °C to 60 °C with batteries removed; 5% to 80% R.H. non-condensing

0 to 2000 m

IEC 61010-1:2001/EN61010-1:2001 (2nd Edition), CAN/CSA-C22.2 No. 61010-1-04,

ANSI/UL 61010-1:2004, CAT II 150 V Overvoltage Protection, Pollution Degree 2

IEC61326-2-1:2005/EN61326-2-1:2006, ICES-001:2004, AS/NZS CISPR11:2004

Input: 0.15 x (specified accuracy)/°C (from 0 °C to 18 °C or 28 °C to 40 °C)

Output: ±(50 ppm output + 0.5 digit)/°C

192 mm x 90 mm x 54 mm

0.98 kg with holster and batteries

One-year calibration cycle recommended

3 years for main unit

3 months for standard accessories unless otherwise specified

Amplicon.com

IT and Instrumentation for industry

Sales: +44 (0) 1273 570 220 Website: www.amplicon.com Email: [email protected]

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Download PDF

advertisement