Extech Instruments SD200 3-Channel Temperature Datalogger User manual
Extech Instruments SD200 is a 3-Channel Datalogging Thermometer that simultaneously displays temperature readings from one to three type k thermocouple probes with an LCD screen. It has a wide temperature range of -58 to 2372°F / -50 to 1300°C and a resolution of 0.1°F/°C. The device can store data on an SD card in Excel® format for easy transfer to a PC, and it also allows for real-time data streaming to a PC via the RS232 output.
Oglas
Oglas
USER MANUAL
3 Channel Datalogging Thermometer
Model SD200
Introduction
Congratulations on your purchase of the Extech SD200 3-Channel Temperature Datalogger. This meter displays and stores temperature readings from one to three type k thermocouple temperature probes. Data is stored on a SD card for transfer to a PC. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Please visit the Extech
Instruments website ( www.extech.com
) to check for the latest version of this User Guide. Extech
Instruments is an ISO-9001 certified company.
Features
Triple LCD simultaneously displays 3 Type-K Temperature channels
Datalogger date/time stamps and stores readings on an SD card in Excel® format for easy transfer to a PC
Selectable data sampling rate: 5, 10, 30, 60, 120, 300, 600 seconds
Long battery life or AC adaptor
-Temperature range of -58 to 2372°F / -50 to 1300°C with 0.1°F/°C resolution
2 SD200-en-GB_v2.0 8/18
Product Description
1. T1 Display
2. T2 Display
3. T3 Display
4. LOG (ENTER) button
5. SET button
6. Temperature Units icon
7.
▲(TIME) button
8. ▼ button
9. AC adaptor socket
10. Reset button
11. RS-232 output
12. SD memory card socket
13. T1 thermocouple input
14. T2 Thermocouple input
15. T3 thermocouple input
Note: Battery Compartment and Tilt Stand are located on the back of the meter.
3 SD200-en-GB_v2.0 8/18
Operation
Battery Low Warning, Installation and Replacement
1. When the icon appears in the display the batteries are weak and should be replaced.
However, in-spec. measurements may still be made for several hours, after low battery indicator appears, before the instrument becomes inaccurate.
2. To replace or install batteries, remove the Philips head screw that secures the rear battery cover and lift off the cover.
3. Replace the six AAA batteries (use alkaline heavy duty type), observing polarity.
4. Replace and secure the cover.
Connecting thermocouples
1. Insert the Type K thermocouple mini-connector plug into the mini-connector socket on the bottom of the datalogger observing polarity (wide blade/narrow blade)
2. Up to three probes may be installed.
3. The temperature for inserted probes will be displayed T1, T2, T3 from top to bottom.
4. “-------“ will be displayed for open or unused inputs.
Datalogging
1. Open the left side door and insert a formatted SD card
Note: The SD card should be at least 1GB in capacity.
Note: Do not use memory cards formatted by other meters or cameras. Use the SD card formatting procedure under the advanced features section of this manual to correctly format the card.
Note: The internal clock must be set to the correct time. See the advanced features section of this manual to set the clock.
Note: The default data structure uses a decimal point “.” as the numeric decimal indicator. See the advanced features section of this manual to change this to a comma “,”.
Note: If the SD memory card is not installed, “ EMPTY ” will appear in the display.
Note: If the temperature probe is not installed, random numbers may appear in the data file.
Note: If the temperature probe is not installed, after data download the temperature units symbol may not appear in the display.
Note: Displayed error messages:
CH-
CArd
The memory card is full or there is a problem with the card
LobAt
The battery is low and datalogging is disabled no
CArd
The SD card is not inserted
2. Press the LOGGER button for >2 seconds to begin logging. “DATALOGGER” will appear in the display (between T2 and T3 displays) and the meter will beep each time the data is recorded (if the beeper is enabled).
4 SD200-en-GB_v2.0 8/18
3.
To stop datalogging, press the LOGGER button for >2 seconds. “DATALOGGER” will change to “DATA” and the meter will count down through the recorded data.
NOTE: To avoid corrupting any data, do not remove the memory card without properly ending the record function.
Time/Date/Sample Rate Check
Press and Hold the TIME button for >2 seconds and the display will cycle through the date, time and sample rate information.
SD Card Data Structure
1. When the SD card if first inserted into the datalogger the folder TMC01 is created.
2. The first datalogging session will then create a file TMC01001.XLS. All data will be saved to this file until the number of rows reaches 30,000.
3. After 30, 000 rows a new file, TMC010 02 .XLS is created. This is repeated every 30,000 rows until TMC010 99 .XLS. At this point a new folder, TMC 02 is created and the process is repeated. TMC 10 is the final folder.
Transferring Data to a PC
1. Remove the memory card from the datalogger and plug it into the SD card slot on the PC.
2. Launch Excel and open the data file on the memory card. The file will appear similar to the figure below.
5 SD200-en-GB_v2.0 8/18
Advanced Settings
The SET function is used to:
Format the SD memory card
Set the date and time
Set the sampling time
Set the beeper sound ON/OFF
Set the SD card Decimal character
Select the Temperature units
Set the RS232 data output ON/OFF
1. Press and Hold the SET button for >2 seconds to enter the setting mode. The first function (Sd
F) will appear in the display. Press the SET button to step through the seven functions. Use the
▲ and ▼ buttons to adjust the selected function. Use the “LOGGER” button to step through fields within a function. In the SET mode, if no button is pressed within 5 seconds the logger will revert back to the standard mode.
2. Sd F – Format the SD card. Press the ▲ button to select yES or no. For yES, press the Enter button. When yES and Ent appear, press the Enter key again to format the card and erase all existing data. The screen will display a flashing yEs and ESC while the memory is being erased and formatted.
3. dAtE
– Set the date and time. Press the ▲ or ▼ buttons to adjust the selected (blinking) field.
Press the Enter button to store the value and to step through the various fields.
4. SP-t – Set the sample rate. Press the ▲ button to select the desired sample rate and press
Enter to store the selection. The selections are: 5, 10, 30, 60, 120, 300, 600 seconds and
AUTO. In AUTO, the data will be stored every time there is a temperature change of >1 degree.
5. bEEP - Set the beeper ON or OFF. Press the ▲ button to select ON or OFF and press Enter to store the selection.
6. dEC - Set the SD card Decimal character. Press the ▲ button to select USA (decimal) or Euro
(comma) and press Enter to store the selection.
7. t-CF - Set the Temp. unit to °F or °C and press Enter to store the selection.
8. rS232 - Set the RS232 data output ON/OFF. Press the
▲ button to select ON or off and press
Enter to store the selection.
9. ESC – Exit the setting mode. Press the SET button to return to normal operation.
System RESET
If a condition appears where the CPU does not respond to keystrokes or the meter seems frozen, press the RESET button on the side of the datalogger (use a paper clip or similar pointed object) to return the meter to a working state.
RS232 Interface
For streaming of data to a PC via the RS232 Output jack, the optional 407001-USB kit (RS232 to USB cable and driver CD) along with the 407001 software (available free at www.extech.com) are required.
6 SD200-en-GB_v2.0 8/18
Specifications
Display
Temperature probes
Memory Card
Datalogger Sampling Time
Temperature Compensation
Display update rate
Data Output
Operating Temperature
Operating Humidity
Power Supply
Battery life
60 mm x 50 mm ( 2.4 x 2.0”) LCD
Type K thermocouple
SD memory card, 1 GB to 16 GB.
5/10/30/60/120/300/600 seconds or Auto.
Automatic
Approx. 1 second
RS 232
0 to 50 °C (32 to 122°F)
Less than 85% R.H.
6 AAA (UM4) Alkaline or heavy duty 1.5 V batteries or
9V AC adaptor.
Dependent on sample rate, for new alkaline batteries and 60 second sampling time, > one month is typical. Fast sample rates will significantly reduce battery life.
199 g /0.44 LB
132 x 80 x 32 mm ( 5.2 x 3.1 x 1.3“)
Weight
Dimension
Type K Thermometer
Range Resolution Accuracy (of reading)
-50.0 to 1300.0 °C
-50.1 to -100.0°C
0.1°C
±( 0.5 % + 0.5°C)
±( 0.5 % + 1°C )
-58.0 to 2372.0°F ±( 0.5 % + 1°F )
0.1°F
-58.1 to -148.0°F ±( 0.5 % + 1.8°F )
Note: Above specification tests under the environment RF Field Strength less than 3 V/M & frequency less than 30 MHz only.
Never dispose of used batteries or rechargeable batteries in household waste.
As consumers, users are legally required to take used batteries to appropriate collection sites, the retail store where the batteries were purchased, or wherever batteries are sold.
Disposal: Do not dispose of this instrument in household waste. The user is obligated to take end-of-life devices to a designated collection point for the disposal of electrical and electronic equipment.
Other Battery Safety Reminders
Never dispose of batteries in a fire. Batteries may explode or leak.