Mettler Toledo Transmitters M300 Series Operating Instruction

Add to My manuals
122 Pages

advertisement

Mettler Toledo Transmitters M300 Series Operating Instruction | Manualzz

Operation Manual

Transmitter M300

– pH/Redox, Dissolved Oxygen,

Conductivity/Resistivity, Dissolved Ozone, single channel version

– Cond/Cond dual channel version

– Multiparameter dual channel version for analog sensors

– Multiparameter single and dual channel version for ISM sensors

Transmitter M300

52 121 389

Transmitter M300 2

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

Operation Manual

Transmitter M300

3

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

Content

1 Introduction ______________________________________________________________________________________________ 9

2 Safety instructions________________________________________________________________________________________ 10

2.1 Definition of equipment and documentation symbols and designations __________________________________________ 10

2.2 Correct disposal of the unit ____________________________________________________________________________ 11

3 Unit overview____________________________________________________________________________________________ 12

3.1 Overview 1/4DIN ____________________________________________________________________________________ 12

3.2 Overview 1/2DIN ____________________________________________________________________________________ 12

3.3 Control/Navigation Keys _______________________________________________________________________________ 13

3.3.1 Menu Structure _______________________________________________________________________________ 13

3.3.2 Navigation keys _______________________________________________________________________________ 13

3.3.2.1 Navigating the menu tree _______________________________________________________________ 13

3.3.2.2 Escape _____________________________________________________________________________ 14

3.3.2.3 Enter _______________________________________________________________________________ 14

3.3.2.4 Menu _______________________________________________________________________________ 14

3.3.2.5 Calibration mode _____________________________________________________________________ 14

3.3.2.6 Info mode ___________________________________________________________________________ 14

3.3.3 Navigation of data entry fields ____________________________________________________________________ 14

3.3.4 Entry of data values, selection of data entry options ___________________________________________________ 14

3.3.5 Navigation with u in Display _____________________________________________________________________ 15

3.3.6 ”Save changes” dialog _________________________________________________________________________ 15

3.3.7 Security Passwords ____________________________________________________________________________ 15

3.4 Display ____________________________________________________________________________________________ 15

4 Installation instruction ____________________________________________________________________________________ 16

4.1 Unpacking and inspection of equipment __________________________________________________________________ 16

4.1.1 Panel cutout dimensional information – 1/4DIN models ________________________________________________ 16

4.1.2 Installation procedure – 1/4DIN models ____________________________________________________________ 17

4.1.3 Panel cutout dimensional information – 1/2DIN models ________________________________________________ 18

4.1.4 Installation procedure – 1/2DIN models ____________________________________________________________ 19

4.1.5 Assembly – 1/2DIN version ______________________________________________________________________ 19

4.1.6 1/2DIN version – Dimension drawings _____________________________________________________________ 20

4.1.7 1/2DIN version – Pipe mounting __________________________________________________________________ 20

4.1.8 1/4DIN version – Dimension drawings _____________________________________________________________ 21

4.2 Connection of power supply ____________________________________________________________________________ 22

4.2.1 1/4DIN housing (panel mount) ___________________________________________________________________ 22

4.2.2 1/2DIN housing (wall mount) ____________________________________________________________________ 23

4.3 Connector PIN definition _______________________________________________________________________________ 24

4.3.1 TB1 and TB2 for 1/2DIN and 1/4DIN versions _______________________________________________________ 24

4.3.2 TB3 and TB4* for 1/2DIN and 1/4DIN versions – Conductivity Sensors ____________________________________ 24

4.3.3 TB3 and TB4* for 1/2DIN and 1/4DIN versions – pH/ORP Sensors _______________________________________ 25

4.3.4 TB3 and TB4* for 1/2DIN and 1/4DIN versions –

Dissolved Oxygen/Dissolved Ozone Sensors (except 58 037 221) ______________________________________ 25

4.3.5 TB3 and TB4* for 1/2DIN and 1/4DIN versions –

Dissolved Oxygen Sensor 58 037 221 only (Thornton Models only) _____________________________________ 26

4.3.6 TB3/TB4* – ISM (digital) Sensors for pH, Conductivity and Dissolved Oxygen _______________________________ 26

4.4 Connection of Analogue Sensor for pH/ORP ________________________________________________________________ 27

4.4.1 Connecting the Sensor to the VP Cable _____________________________________________________________ 27

4.4.2 VP Cable Assignment __________________________________________________________________________ 28

4.4.3 Typical Wiring (Using TB3/TB4) __________________________________________________________________ 29

4.4.3.1 Example 1 __________________________________________________________________________ 29

4.4.3.2 Example 2 __________________________________________________________________________ 30

4.4.4 Example 3 ___________________________________________________________________________________ 31

4.4.4.1 Example 4 __________________________________________________________________________ 32

4.5 Connection of Analogue Sensor for Dissolved Oxygen/Dissolved Ozone (except 58 037 221) ________________________ 33

4.5.1 Connecting the Sensor to the VP Cable _____________________________________________________________ 33

4.5.2 Typical Wiring using TB3/TB4 ____________________________________________________________________ 34

4.6 Connection of Analogue Sensor for Dissolved Oxygen 58 037 221 _____________________________________________ 35

4.7 Connection of ISM Sensor ______________________________________________________________________________ 35

4.7.1 Connection of ISM Sensor for pH, Cond 4-e and Dissolved Oxygen _______________________________________ 35

4.7.2 AK9 Cable Assignment _________________________________________________________________________ 36

4.7.3 Connection of ISM Sensor for Cond 2-e (Thornton Models only) _________________________________________ 36

4.7.4 Cable Assignment of ISM Sensor for Cond 2-e (Thornton Models only) ____________________________________ 36

5 Placing transmitter in, or out, of service _____________________________________________________________________ 37

5.1 Placing transmitter in service ___________________________________________________________________________ 37

5.2 Placing transmitter out of service _______________________________________________________________________ 37

4

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 5

6 Quick Setup _____________________________________________________________________________________________ 38

7 Sensor Calibration _______________________________________________________________________________________ 39

7.1 Enter Calibration Mode ________________________________________________________________________________ 39

7.2 Conductivity/Resistivity Calibration ______________________________________________________________________ 40

7.2.1 One-point Sensor Calibration ____________________________________________________________________ 41

7.2.2 Two-point Sensor Calibration (4-electrode sensors only) _______________________________________________ 41

7.2.3 Process Calibration ___________________________________________________________________________ 42

7.3 Oxygen Calibration ___________________________________________________________________________________ 43

7.3.1 One-Point Sensor Calibration ____________________________________________________________________ 43

7.3.1.1 Auto mode __________________________________________________________________________ 43

7.3.1.2 Manual mode ________________________________________________________________________ 44

7.3.2 Process Calibration ____________________________________________________________________________ 44

7.4 Ozone Calibration (Thornton Models only) _________________________________________________________________ 44

7.4.1 One-Point Sensor Calibration ____________________________________________________________________ 45

7.5 pH Calibration ______________________________________________________________________________________ 45

7.5.1 One Point Calibration __________________________________________________________________________ 45

7.5.1.1 Auto Mode __________________________________________________________________________ 46

7.5.1.2 Manual Mode ________________________________________________________________________ 46

7.5.2 Two Point Calibration __________________________________________________________________________ 46

7.5.2.1 Auto Mode __________________________________________________________________________ 47

7.5.2.2 Manual Mode ________________________________________________________________________ 47

7.5.3 Process Calibration ___________________________________________________________________________ 48

7.5.4 mV Calibration (not on ISM Version) _______________________________________________________________ 48

7.5.5 ORP Calibration (only on ISM Version) ____________________________________________________________ 49

7.6 Sensor Temperature Calibration (not on ISM Version) ________________________________________________________ 49

7.6.1 One-Point Sensor Temperature Calibration (not on ISM Version)__________________________________________ 49

7.6.2 Two-Point Sensor Temperature Calibration (not on ISM version) __________________________________________ 50

7.7 Edit Sensor Calibration Constants (not on ISM version) _______________________________________________________ 50

7.8 Sensor Verification ___________________________________________________________________________________ 51

8 Configuration ____________________________________________________________________________________________ 52

8.1 Enter Configuration Mode ______________________________________________________________________________ 52

8.2 Measurement _______________________________________________________________________________________ 52

8.2.1 Channel Setup ________________________________________________________________________________ 52

8.2.2 Derived Measurements (Thornton models only) ______________________________________________________ 53

8.2.2.1 % Rejection measurement ______________________________________________________________ 54

8.2.2.2 Calculated pH (Power plant applications only) ______________________________________________ 54

8.2.2.3 Calculated CO

2

(Power plant applications only) ______________________________________________ 55

8.2.3 Temperature Source (not on ISM version) ___________________________________________________________ 55

8.2.4 Parameter Related Settings ______________________________________________________________________ 56

8.2.4.1 Conductivity/Resistivity Temperature Compensation ___________________________________________ 57

8.2.4.2 pH/ORP Parameters ___________________________________________________________________ 58

8.2.4.3 Dissolved Oxygen Parameters ___________________________________________________________ 59

8.2.5 Set Averaging ________________________________________________________________________________ 60

8.3 Analog Outputs ______________________________________________________________________________________ 60

8.4 Setpoints __________________________________________________________________________________________ 62

8.5 Alarm/Clean ________________________________________________________________________________________ 65

8.5.1 Alarm _______________________________________________________________________________________ 65

8.5.2 Clean _______________________________________________________________________________________ 66

8.6 Display ____________________________________________________________________________________________ 66

8.6.1 Measurement _________________________________________________________________________________ 67

8.6.2 Resolution ___________________________________________________________________________________ 67

8.6.3 Backlight ____________________________________________________________________________________ 67

8.6.4 Name ______________________________________________________________________________________ 68

8.7 Hold Analog Outputs _________________________________________________________________________________ 68

9 System _________________________________________________________________________________________________ 69

9.1 Set Language _______________________________________________________________________________________ 69

9.2 USB ______________________________________________________________________________________________ 69

9.3 Passwords _________________________________________________________________________________________ 70

9.3.1 Changing Passwords __________________________________________________________________________ 70

9.3.2 Configuring Menu Access for Operator _____________________________________________________________ 70

9.4 Set/Clear Lockout ____________________________________________________________________________________ 70

9.5 Reset _____________________________________________________________________________________________ 71

9.5.1 Reset System _________________________________________________________________________________ 71

9.5.2 Reset Meter Calibration (not on ISM version) ________________________________________________________ 71

9.5.3 Reset Analog Calibration ________________________________________________________________________ 72

9.5.4 Reset Sensor Calibration Data to Factory Settings _____________________________________________________ 72

9.5.5 Reset Calibration Data of Sensor Electronic to Factory Settings ___________________________________________ 72

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 6

10 PID Setup _______________________________________________________________________________________________ 73

10.1 Enter PID Setup _____________________________________________________________________________________ 74

10.2 PID Auto/Manual ____________________________________________________________________________________ 75

10.3 Mode _____________________________________________________________________________________________ 75

10.3.1 PID Mode ___________________________________________________________________________________ 75

10.4 Tune Parameters _____________________________________________________________________________________ 76

10.4.1 PID Assignment & Tuning _______________________________________________________________________ 77

10.4.2 Setpoint & Deadband __________________________________________________________________________ 77

10.4.3 Proportional Limits ____________________________________________________________________________ 77

10.4.4 Corner Points _________________________________________________________________________________ 77

10.5 PID Display ________________________________________________________________________________________ 77

11 Service _________________________________________________________________________________________________ 78

11.1 Enter Service Menu ___________________________________________________________________________________ 78

11.2 Diagnostics ________________________________________________________________________________________ 78

11.2.1 Model/Software Revision ________________________________________________________________________ 78

11.2.2 Digital Input __________________________________________________________________________________ 79

11.2.3 Display _____________________________________________________________________________________ 79

11.2.4 Keypad _____________________________________________________________________________________ 79

11.2.5 Memory _____________________________________________________________________________________ 79

11.2.6 Set Relay ____________________________________________________________________________________ 80

11.2.7 Read Relays _________________________________________________________________________________ 80

11.2.8 Set Analog Outputs ____________________________________________________________________________ 80

11.2.9 Read Analog Outputs ___________________________________________________________________________ 80

11.3 Calibrate ___________________________________________________________________________________________ 81

11.3.1 Calibrate Meter (not on ISM version) _______________________________________________________________ 81

11.3.1.1 Resistance __________________________________________________________________________ 81

11.3.1.2 Temperature _________________________________________________________________________ 82

11.3.1.3 Current _____________________________________________________________________________ 83

11.3.1.4 Voltage _____________________________________________________________________________ 83

11.3.1.5 Rg Diagnostic ________________________________________________________________________ 84

11.3.1.6 Rr Diagnostics _______________________________________________________________________ 84

11.3.2 Calibrate Analog ______________________________________________________________________________ 85

11.3.3 Calibrate Unlock ______________________________________________________________________________ 85

11.4 Tech Service ________________________________________________________________________________________ 85

12 Info ____________________________________________________________________________________________________ 86

12.1 Info Menu __________________________________________________________________________________________ 86

12.2 Messages __________________________________________________________________________________________ 86

12.3 Calibration Data _____________________________________________________________________________________ 86

12.4 Model/Software Revision ______________________________________________________________________________ 87

12.5 ISM Sensor Info (ISM version only) ______________________________________________________________________ 87

13 Maintenance ____________________________________________________________________________________________ 88

13.1 For Technical Support _________________________________________________________________________________ 88

13.2 Front Panel Cleaning _________________________________________________________________________________ 88

14 Troubleshooting __________________________________________________________________________________________ 89

14.1 Changing the Fuse ___________________________________________________________________________________ 89

14.2 pH Error Messages / Warning- and Alarm List _______________________________________________________________ 90

14.2.1 pH sensors execpt dual membrane pH electrodes ___________________________________________________ 90

14.2.2 Dual membrane pH electrodes (pH/pNa) ___________________________________________________________ 91

14.2.3 ORP messages ______________________________________________________________________________ 91

14.3 O2 Error Messages / Warning- and Alarm List ______________________________________________________________ 92

14.4 Cond Error Messages / Warning- and Alarm List ____________________________________________________________ 92

14.5 O

14.6 O

2

2

(I) Error Messages / Warning- and Alarm List (Thornton Models only) _________________________________________ 92

(V) Error Messages / Warning- and Alarm List (Thornton Models only) _________________________________________ 93

14.7 Ozone Error Messages / Warning- and Alarm List (Thornton Models only) _________________________________________ 93

14.8 Warning- and Alarm indication on the display _____________________________________________________________ 93

14.8.1 Warning indication ____________________________________________________________________________ 93

14.8.2 Alarm indication ______________________________________________________________________________ 94

15 Accessories and Spare Parts _______________________________________________________________________________ 95

16 Specifications ___________________________________________________________________________________________ 96

16.1 General specifications ________________________________________________________________________________ 96

16.2 Electrical specifications for

1/2DIN and 1/4DIN versions ___________________________________________________________________________ 98

16.3 Mechanical specifications for 1/4DIN version ______________________________________________________________ 98

16.4 Mechanical specifications for 1/2DIN version ______________________________________________________________ 99

16.5 Environmental specifications for 1/2DIN and 1/4DIN versions __________________________________________________ 99

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 7

17 Default tables __________________________________________________________________________________________ 100

17.1 M300 ISM (Single-Channel Instruments) _________________________________________________________________ 100

17.2 M300 ISM (Dual-Channel Instruments) __________________________________________________________________ 102

17.3 M300 Conductivity (Single-Channel Instruments) __________________________________________________________ 104

17.4 M300 O

2

(Single-Channel Instruments) __________________________________________________________________ 105

17.5 M300 pH (Single-Channel Instruments) _________________________________________________________________ 107

17.6 M300 Multiparameter (Dual-Channel Instruments) _________________________________________________________ 109

17.7 M300 Conductivity (Dual-Channel Instruments, Thornton Models only) _________________________________________ 112

18 Warranty ______________________________________________________________________________________________ 114

19 Certificate _____________________________________________________________________________________________ 115

20 Buffer tables ___________________________________________________________________________________________ 116

20.1 Standard pH buffers _________________________________________________________________________________ 116

20.1.1 Mettler-9 ___________________________________________________________________________________ 116

20.1.2 Mettler-10 __________________________________________________________________________________ 117

20.1.3 NIST Technical Buffers _________________________________________________________________________ 117

20.1.4 NIST standard buffers (DIN and JIS 19266: 2000–01) _______________________________________________ 118

20.1.5 Hach buffers ________________________________________________________________________________ 118

20.1.6 Ciba (94) buffers ____________________________________________________________________________ 119

20.1.7 Merck Titrisole, Riedel-de-Haën Fixanale __________________________________________________________ 119

20.1.8 WTW buffers ________________________________________________________________________________ 120

20.1.9 JIS Z 8802 buffers ____________________________________________________________________________ 120

20.2 Dual membrane pH electrode buffers ____________________________________________________________________ 121

20.2.1 Mettler-pH/pNa buffers (Na+ 3.9M) ______________________________________________________________ 121

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 8

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

1 Introduction

Statement of Intended Use – The M300 Multiparameter transmitter is a single- or dual channel online process instrument for measuring various properties of fluids. These include Conductivity/

Resistivity, Dissolved Oxygen, Dissolved Ozone and pH/ORP. It will interface with a variety of different Mettler-Toledo sensors, which connect to the transmitter using cables of varied lengths.

Depending on the type, the transmitter can handle conventional analog sensors or the future oriented ISM (Intelligent Sensor Management) sensors.

A large four line backlit Liquid Crystal Display conveys measuring data and setup information.

The menu structure allows the operator to modify all operational parameters by using keys on the front panel. A menu-lockout feature, with password protection, is available to prevent the unauthorized use of the meter. The M300 Multiparameter transmitter can be configured to use its 2

(4 on dual channel version) analog and/or 4 (6 on dual channel version) relay outputs for process control.

The M300 Multiparameter transmitter is equipped with a USB communication interface. This interface provides real-time data output and complete instrument configuration capabilities for central monitoring via Personal Computer (PC).

This manual applies for all available M300 transmitters as follows:

– Single parameter and single channel version for pH/ORP, dissolved Oxygen, Conductivity/

Resistivity, and dissolved Ozone

– Mulitparameter dual channel version for analog sensors

– Mulitparameter single and dual channel version for ISM sensors

– Dual channel Cond/Cond version for analog sensors

M300 parameter fit guide

M300 Thornton models

Designation

M300 ISM 1-channel 1/4DIN

M300 ISM 1-channel 1/2DIN

M300 ISM 2-channel 1/4DIN

Order no.

58 000 301

58 000 311

58 000 302

Analog Sensors

M300 ISM 2-channel 1/2DIN

M300 pH 1-channel 1/4DIN

58 000 312

58 001 303 pH

M300 pH 1-channel 1/2DIN 58 001 313 pH

M300 Cond 1-channel 1/4DIN 58 002 301 Cond

M300 Cond 1-channel 1/2DIN 58 002 311 Cond

M300 Cond 2-channel 1/4DIN 58 001 304 Cond

M300 Cond 2-channel 1/2DIN 58 001 314 Cond

M300 Multi 2-channel 1/4DIN 58 001 306 pH, Cond, DO ppm*,

DO ppb*, O3*

M300 Multi 2-channel 1/2DIN 58 001 316 pH, Cond, DO ppm*,

DO ppb*, O3*

* THORNTON sensors

ISM Sensors pH, DO*, Cond pH, DO*, Cond pH, DO*, Cond pH, DO*, Cond

9

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 10

M300 Ingold models

Designation

M300 ISM 1-channel 1/4DIN

M300 ISM 1-channel 1/2DIN

M300 ISM 2-channel 1/4DIN

Order no.

52 121 354

52 121 355

52 121 356

Analog Sensors

M300 ISM 2-channel 1/2DIN

M300 pH 1-channel 1/4DIN

52 121 357

52 121 286 pH

M300 pH 1-channel 1/2DIN 52 121 289 pH

M300 Cond 1-channel 1/4DIN 52 121 288 Cond

M300 Cond 1-channel 1/2DIN 52 121 291 Cond

M300 O2 1-channel 1/4DIN

M300 O2 1-channel 1/2DIN

M300 Multi 2-channel 1/4DIN

M300 Multi 2-channel 1/2DIN

52 121 287 DO**

52 121 290 DO**

52 121 292

52 121 293 pH, Cond, DO** pH, Cond, DO**

** INGOLD sensors

ISM Sensors pH, DO**, Cond 4-e pH, DO**, Cond 4-e pH, DO**, Cond 4-e pH, DO**, Cond 4-e

The print screen images in this manual have a general explaining character and can differ from the real display in your transmitter.

This description corresponds to the firmware release, version 1.4 of the M300 ISM transmitter

(resp. version 1.1 of THORNTON M300 ISM transmitter) and version 1.6 for all other M300 transmitters. Changes are taking place constantly, without prior notification

2 Safety instructions

This manual includes safety information with the following designations and formats.

2.1 Definition of equipment and documentation symbols and designations

a

WARNING: POTENTIAL FOR PERSONAL INJURY.

a

CAUTION: possible instrument damage or malfunction.

h

NOTE: Important operating information.

a

On the transmitter or in this manual text indicates: Caution and/or other possible hazard including risk of electric shock (refer to accompanying documents).

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

h h h

Transmitter M300 11

The following is a list of general safety instructions and warnings. Failure to adhere to these instructions can result in damage to the equipment and/or personal injury to the operator.

– The M300 Transmitter should be installed and operated only by personnel familiar with the transmitter and who are qualified for such work.

– The M300 Transmitter must only be operated under the specified operating conditions

(see section 16).

– Repair of the M300 Transmitter must be performed by authorized, trained personnel only.

– With the exception of routine maintenance, cleaning procedures or fuse replacement, as described in this manual, the M300 Transmitter must not be tampered with or altered in any manner.

– Mettler-Toledo accepts no responsibility for damage caused by unauthorized modifications to the transmitter.

– Follow all warnings, cautions, and instructions indicated on and supplied with this product.

– Install equipment as specified in this instruction manual. Follow appropriate local and national codes.

– Protective covers must be in place at all times during normal operation.

– If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired.

WARNINGS:

Installation of cable connections and servicing of this product require access to shock hazard voltage levels.

Main power and relay contacts wired to separate power source must be disconnected before servicing.

Switch or circuit breaker shall be in close proximity to the equipment and within easy reach of the OPERATOR; it shall be marked as the disconnecting device for the equipment.

Main power must employ a switch or circuit breaker as the disconnecting device for the equipment.

Electrical installation must be in accordance with the National Electrical Code and/or any other applicable national or local codes.

NOTE! RELAY CONTROL ACTION: the M300 transmitter relays will always de-energize on loss of power, equivalent to normal state, regardless of relay state setting for powered operation. Configure any control system using these relays with fail-safe logic accordingly.

NOTE! PROCESS UPSETS: Because process and safety conditions may depend on consistent operation of this transmitter, provide appropriate means to maintain operation during sensor cleaning, replacement or sensor or instrument calibration.

NOTE: This is a 4-wire-product with an active 4–20 mA analog output.

Please do not supply to Pin1–Pin6 of TB2.

2.2 Correct disposal of the unit

When the transmitter is finally removed from service, observe all local environmental regulations for proper disposal.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

3 Unit overview

M300 models are available in both a 1/4DIN and 1/2DIN case size. The 1/4DIN is a panelmount only design and the 1/2DIN models provide an integral IP65 housing for wall-, or pipemount.

12

3.1 Overview 1/4DIN

1

4.01

[102]

8 M300

4

3

5

4.01

[102]

6

7

ESC

Menu Cal

2

Info Enter

1: Hard Polycarbonate case

2: Five Tactile-Feedback Navigation Keys

3: Four-line LCD Display

4: Power Supply Terminals

3.2

1

Overview 1/2DIN

5.90

[150]

M300

3

5.90

[150]

Menu

ESC

Cal

2

Info Enter

1: Hard Polycarbonate case

2: Five Tactile-Feedback Navigation Keys

3: Four-line LCD Display

4: Power Supply Terminals

5: USB Interface Port

6: Relay Output Terminals

7: Analog Output/Digital Input Terminals

8: Sensor Input Terminals

5

7

6

8

4

5: USB Interface Port

6: Relay Output Terminals

7: Analog Output/Digital Input Terminals

8: Sensor Input Terminals

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

3.3 Control/Navigation Keys

3.3.1 Menu Structure

Below is the structure of the M300 menu tree:

Quick Setup

Channel Setup

Configure

Measurement

System

Set Language

PID Setup

PID A/M

Menu

Service

Measurement

Mode M300

Cal

Messages

Info

ISM Sensor

Info*

Calibration Data

Model/Software

Revision

Diagnostics

Analog Outputs Analog Outputs USB Tune Parameters Calibrate

Set Points Set Points

Alarm/Clean

Passwords

Set/Clear

Lockout

Reset

Mode

PID

Display Setup

Tech Service

Display

Hold Outputs

* on ISM version only

13 h

3.3.2 Navigation keys

ESC

Menu Cal Info Enter

3.3.2.1 Navigating the menu tree

Enter the desired main Menu branch with the b c

or m

keys. Use the m

and

.

keys to navigate through the selected Menu branch.

NOTE: In order to back up one menu page, without escaping to the measurement mode, move the cursor under the UP Arrow character ( u ) at the bottom right of the display screen and press [Enter].

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 h

3.3.2.2 Escape

Press the b and c key simultaneously (escape) to return to the Measurement mode.

3.3.2.3 Enter

Use the e key to confirm action or selections.

3.3.2.4 Menu

Press the b key to access the main Menu.

14

3.3.2.5 Calibration mode

Press the c key to enter Calibration Mode.

3.3.2.6 Info mode

Press the .

key to enter Info Mode

3.3.3 Navigation of data entry fields

Use the c key to navigate forward or the data entry fields of the display.

b key to navigate backwards within the changeable

3.3.4 Entry of data values, selection of data entry options

Use the m key to increase or the .

key to decrease a digit. Use the same keys to navigate within a selection of values or options of a data entry field.

NOTE: Some screens require configuring multiple values via the same data field (ex: configuring multiple setpoints). Be sure to use the c or b key to return to the primary field and the m or

.

key to toggle between all configuration options before entering to the next display screen.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 h h

3.3.5 Navigation with

u

in Display

If a u is displayed on the bottom right hand corner of the display, you can use the c or the b key to navigate to it. If you click [ENTER] you will navigate backwards through the menu

(go back one screen). This can be a very useful option to move back up the menu tree without having to exit into the measuring mode and re-enter the menu.

15

3.3.6 ”Save changes” dialog

Three options are possible for the ”Save changes” dialog: Yes & Exit (Save changes and exit to measuring mode), ”Yes & u ” (Save changes and go back one screen) and ”No & Exit” (Don’t save changes and exit to measuring mode). The ”Yes & u ” option is very useful if you want to continue configuring without having to re-enter the menu.

3.3.7 Security Passwords

The M300 transmitter allows a security lock-out of various menus. If the security lock-out feature of the transmitter has been enabled, a security password must be entered to allow access to the menu. See section 9.3 for more information.

3.4 Display

NOTE: In the event of an alarm or other error condition the M300 Transmitter will display a flashing a in the upper right corner of the display. This symbol will remain until the condition that caused it has been cleared.

NOTE: During calibrations, clean, Digital In with Analog Output/Relay/USB in Hold state, a flashing H will appear in the upper left corner of the display. This symbol will remain for

20 seconds until after the calibration or clean is completed. This symbol will also disappear when Digital In is deactivated.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4 Installation instruction

16

4.1 Unpacking and inspection of equipment

Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions.

Do not discard the box.

If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present.

If items are missing, notify Mettler-Toledo immediately.

4.1.1 Panel cutout dimensional information –

1/4DIN models

1/4DIN Model transmitters are designed for panel-mount installation only. Each transmitter is supplied with mounting hardware to provide fast and simple installation to a flat panel or flat enclosure door. To insure a good seal and maintain IP integrity of installation, the panel or door must be flat and have a smooth finish. Hardware consists of:

2 snap-on Mounting brackets

1 mounting gasket seal

Transmitter dimensions and mounting are shown in the figures below.

3.62

[92

–0

+0.02

+0.5

]

3.62

[92

+0.02

+0.5

–0

]

PANEL CUT-OUT

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

MOUNTING

BRACKET

2 PLACES

Transmitter M300

4.1.2 Installation procedure – 1/4DIN models

– Make cutout in panel (see dimensions cutout drawing).

– Be sure surface surrounding cutout is clean, smooth and free of burrs.

– Slide face gasket (supplied with transmitter) around transmitter from the back of the unit.

– Place transmitter into cutout hole. Be sure there are no gaps between the transmitter and panel surface.

– Place the two mounting brackets on either side of the transmitter as shown.

– While holding transmitter firmly into the cutout hole, push the mounting brackets toward the backside of panel.

– Once secure, use a screwdriver to tighten the brackets against the panel. In order to provide

IP65 environmental enclosure rating, the two clamps provided shall be securely tightened to create an adequate seal between the panel enclosure and M300 front face.

– Face gasket will compress between transmitter and panel.

a

CAUTION: Do not over tighten brackets.

4.33

[110]

4.33

[110]

4.01

[102]

3.55

[90] 4.01

[102]

3.55

[90]

5

4

3

2

1

9

8

7

6

3.55

[90]

Power

USB

Fuse

1 2 3 4 5 6 7

1 2 3 4 5 6 7

5

4

3

2

1

9

8

7

6

9

8

7

6

5

4

3

2

1

4.01

[102]

4.01

[102]

3.55

[90]

Power Fuse

1 2 3 4 5 6 7 5

4

3

2

1

9

8

7

6

USB 1 2 3 4 5 6 7

17

4.01

[102]

MAX. 0.118

[3]

FLAT GASKET MOUNTING

BRACKET

2 PLACES

4.01

[102]

4.96

[126]

4.92

[125]

MAX. 0.118

[3]

FLAT GASKET

4.96

[126]

4.92

[125] 4.33

[110]

1.77

[45]

3.54

[90]

4.33

[110]

1.77

[45]

3.54

[90]

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 18

4.1.3 Panel cutout dimensional information –

1/2DIN models

1/2DIN Model transmitters are designed with an integral rear cover for stand-alone wall mount installation.

The unit may also be wall mounted using the integral rear cover. See installation instructions in

Section 4.1.4.

Below are cut-out dimensions required by the 1/2DIN models when mounted within a flat panel or on a flat enclosure door. This surface must be flat and smooth. Textured or rough surfaces are not recommended and may limit the effectiveness of the gasket seal provided.

5.39 +0.02

–0.00

[137 ]

5.39 +0.02

–0.00

[137 +0.5

–0.0

]

PANEL CUT-OUT

Optional hardware accessories are available that allow for panel- or pipe-mount.

Refer to Section 15 “Accessories and Spare Parts” for ordering information.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 19

4.1.4 Installation procedure – 1/2DIN models

General:

– Orient the transmitter so that the cable grips face downward.

– Wiring routed through the cable grips shall be suitable for use in wet locations.

– In order to provide IP65 enclosure ratings, all cable glands must be in place. Each cable gland must be filled using a cable, or suitable Cable Gland Hole Seal.

For Wall Mount:

– Remove rear cover from front housing.

– Start by unscrewing the four screws located on the face of the transmitter, in each corner.

This allows the front cover to swing away from the rear housing.

– Remove the hinge-pin by squeezing the pin from each end.

This allows the front housing to be removed from the rear housing.

– Mount rear housing to wall. Secure mounting kit to the M300 according to the supplied instructions. Attach to wall using appropriate mounting hardware for wall surface. Be sure it is level and securely fastened and the installation adheres to any and all clearance dimensions required for transmitter service and maintenance. Orient the transmitter so that the cable grips are facing downward.

– Replace the front housing to the rear housing. Securely tighten the rear-cover screws to ensure that IP65 enclosure environmental rating is provided. The unit is ready to be wired.

For Pipe Mount:

– Use only manufacturer-supplied components for pipe-mounting the M300 transmitter and install per the supplied instructions. See section 15 for ordering information.

4.1.5 Assembly – 1/2DIN version

1: 3 Pg 13.5 cable glands

2: 2 plastic plugs

3: 4 screws

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

1

2

3

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.1.6 1/2DIN version – Dimension drawings

150 mm/5.9"

116 mm/4.57"

29 mm

1.14"

20

42 mm

1.65"

84 mm

3.31"

35 mm/

1.38"

90 mm/3.54"

80 mm/3.15"

29 mm

1.14"

137 mm (+ 0,5 mm/– 0,0 mm)

5.39" (+ 0.02"/– 0.00")

4.1.7 1/2DIN version – Pipe mounting

40 ... 60 mm

1.57... 2.36

"

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.1.8 1/4DIN version – Dimension drawings

110 mm / 4.33"

102 mm / 4.02"

90 mm / 3.54"

TB4 TB3

Fuse

1 2

Power

92 mm ± 0.5 mm

3.62" ± 0.02"

USB

TB1A

TB1B

TB2

21

102 mm / 4.02"

110 mm / 4.33" 90 mm / 3.54"

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.2 Connection of power supply

All connections to the transmitter are made on the rear panel of all models.

a

Be sure power to all wires is turned off before proceeding with the installation. High voltage may be present on the input power wires and relay wires.

A two-terminal connector on the rear panel of all M300 models is provided for power connection. All M300 models are designed to operate from a 20–30 VDC or a 100 to 240 VAC power source. Refer to specifications for power requirements and ratings and size power wiring accordingly.

The terminal block for power connections is labeled ”Power” on the rear panel of the transmitter.

One terminal is labeled – N for the Neutral wire and the other + L for the Line (or Load) wire. The terminals are suitable for single wires and flexible leads up to 2.5 mm 2 (AWG 14). There is no earth ground terminal on the transmitter. For this reason the internal power wiring within the transmitter is double insulated and the product label designates this using the d symbol.

22

4.2.1 1/4DIN housing (panel mount)

2

1

1: Connection of power supply

2: Terminal for sensors

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.2.2 1/2DIN housing (wall mount)

23

1: Connection of power supply

2: Terminal for sensors

2

1

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.3 Connector PIN definition

24

4.3.1 TB1 and TB2 for 1/2DIN and 1/4DIN versions

1 ⁄

4

DIN

Power connections are labeled – N for Neutral and +L for Line, for 100 to 240 VAC or 20 – 30 VDC.

TB2 for

1 AO1+

1 ⁄

4

DIN

2 AO1–/AO2–

3 AO2+

4 AO3+*

5 AO3–/AO4–*

6 AO4+*

7 DI1+

8 DI1–/DI2–*

9 DI2+*

TB1A

1

1

7

7

TB4

9

1

TB1B

TB3

9

TB2

1

1

9

1 ⁄

2

DIN

TB2 for

1 AO1+

1 ⁄

2

DIN

2 AO1–/AO2–

3 AO2+

4 AO3+*

5 AO3–/AO4–*

6 AO4+*

7 DI1+

8 DI1–/DI2–*

9 DI2+*

TB1 1

1

TB2

9 1

14

TB3

9 1

TB4

9

TB1A for 1 ⁄

4

1 NO2

2 COM2

3 NC2

4 NO6*

5 COM6*

6 NO4

7 COM4

DIN

* Dual channel only

TB1B for 1 ⁄

4

1 NO1

2 COM1

3 NC1

4 NC5*

5 COM5*

6 NO3

7 COM3

DIN

1 NO1

2 COM1

3 NC1

4 NO2

5 COM2

6 NC2

7 COM5*

TB1 for 1 ⁄

2

DIN

8 NC5*

9 COM6*

10 NO6*

11 NO3

12 COM3

13 NO4

14 COM4

NO: normally open (contact open if un-actuated)

NC: normally closed (contact closed if un-actuated)

AO: Analog Output

DI: Digital Input

4.3.2 TB3 and TB4* for 1/2DIN and 1/4DIN versions –

Conductivity Sensors

TB 3 provides access to channel A signal inputs, TB4* provides access to channel B signal inputs.

Conductivity sensors use 58 080 20X or 58 080 25X series cables.

5

6

7

8

9

Pin no.

1

2

3

4

* Only on dual channel version

** Transparent not connected.

Sensor wire Color** white white / blue blue black

– bare shield red green

Function

Cnd inner 1

Cnd outer 1

Cnd inner 2

Cnd outer 2 / Shield not used

RTD ret / GND

RTD sense

RTD

+5V

Terminal 4 and 6 are internally connected, either terminal can be used to connect a wire.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 h h

4.3.3 TB3 and TB4* for 1/2DIN and 1/4DIN versions – pH/ORP Sensors

pH/ORP sensors use 52 300 1XX series VP cables, or 10 001 XX02 series AS9 cables

(ORP only).

7

8

5

6

Pin no.

1

2

3**

4**

9

Sensor wire Color

Coax inner/transparent

Coax shield/red green/yellow, blue

– white green

– grey (no connection)

* Only on dual channel version.

Function

Glass not used

Reference

Solution GND/Shield not used

RTD ret/GND

RTD sense

RTD

+5V

Terminal 4 and 6 are internally connected, either terminal can be used to connect a wire.

NOTE: ** Install Jumper 3 to 4 when used without Solution Ground.

NOTE: For sensors with Pt100 temperature probe, the Pt100 adapter is required.

The Pt100 adapter is enclosed in the package of each transmitter.

25 h

4.3.4 TB3 and TB4* for 1/2DIN and 1/4DIN versions –

Dissolved Oxygen/Dissolved Ozone Sensors

(except 58 037 221)

These sensors use 52 300 1XX series VP cables.

8

9

6

7

Pin no.

1**

2

3**

4**

5

Blue wire not used.

* Only on dual channel version.

Sensor wire Color

Coax Shield/red

– green/yellow

Coax Inner/transparent white, grey

– green

Function not used

Anode not used

Shield/GND

Cathode

Temperature, Guard not used

Temperature

+5V

Terminal 4 and 6 are internally connected, either terminal can be used to connect a wire.

NOTE: ** Install Jumper (supplied) 1 to 3 to 4 when using Thornton Dissolved Oxygen and

Ozone Sensors.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.3.5 TB3 and TB4* for 1/2DIN and 1/4DIN versions –

Dissolved Oxygen Sensor 58 037 221 only

(Thornton Models only)

This sensor uses 58 080 25X series cables.

7

8

5

6

2

3

Pin no.

1

4

9

* Only on dual channel version.

Sensor wire Color white white/blue,

– black, bare shield

Transparent

Red

Green

Blue

Function

Signal

Range

Shield, Ground

Ground

Temperature

Temperature

+5V

Terminal 4 and 6 are internally connected, either terminal can be used to connect a wire.

26

4.3.6 TB3/TB4* –

ISM (digital) Sensors for pH, Conductivity and

Dissolved Oxygen

The wiring of the digital 9 terminal connectors is:

8

9

6

7

4

5

2

3

Pin no.

1

* Only on dual channel version

** Bare wire not connected

*** For Thornton models only

Function

24 VDC

GND (24 VDC)

1-Wire

GND (5 VDC)

GND (5 VDC)

RS485-B

RS485-A

5 VDC

– pH, Oxygen, Cond 4-e

Color

– transparent (cable core) red (shield)

Cond 2-e***

Color**

– white black red blue

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.4 Connection of Analogue Sensor for pH/ORP

4.4.1 Connecting the Sensor to the VP Cable

27

A

B

C

F

D

E

A B C D E F S h NOTE: Cable lengths > 20 m can worsen the response during pH measurement. Be sure to observe the sensor instruction manual.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.4.2 VP Cable Assignment

A transparent

B red

Glass electrode

Ref. electrode

C gray

D blue

E white

T3

Solution ground

T1

F green

S green/yellow

T2

Outer shield

C = 220 nF

T1/T2: Temperature probe for 2-wire connection

T3: Additional connection for temperature probe (3-wire connection)

28

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.4.3 Typical Wiring (Using TB3/TB4)

4.4.3.1 Example 1

pH measurement without Solution Ground.

9

8

7

6

5

2

1

4

3

Jumper

Cable

Temperature probe

29

Pt100 Adapter wiring diagramm for TB3/4

Change M300 setting to Pt100 under

Configuration/Measurement/Temperature Source

2-wire

Pt100

Adapter

9

8

7

4

3

2

1

6

5 h h

Temperature probe

3-wire

Pt100

Adapter

Combination pH electrode

Temperature probe

NOTE: Jumper Terminals 3 and 4.

NOTE: Pt100 adapter required (enclosed) for sensors with Pt100 temperature probe.

For wiring details see page 24.

Wire colors only valid for connection with VP cable; blue and grey not connected.

1: Glass

2: Not used

3: Reference

4: Shield/GND

5: Not used

6: Solution GND/RTD ret

7: Not used

8: RTD

9: Not used

9

8

7

6

5

2

1

4

3

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.4.3.2 Example 2

pH measurement with Solution Ground

4

3

2

1

6

5

9

8

7

Cable h h

Temperature probe

Combination pH electrode with RTD and SG

NOTE: Wire colors only valid for connection with VP cable, grey not connected.

NOTE: Pt100 adapter required (enclosed) for sensors with Pt100 temperature probe.

For wiring details see page 24.

1: Glass

2: Not used

3: Reference

4: Shield/Solution GND

5: Not used

6: GND/RTD ret

7: Not used

8: RTD

9: Not used

30

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.4.4 Example 3

ORP (redox) measurement (temperature optional).

9

8

7

6

5

4

3

2

1

Jumper

Cable

31

Temperature probe

ORP electrode

h h

NOTE: Jumper Terminal 3 and 4.

NOTE: Pt100 adapter required (enclosed) for sensors with Pt100 temperature probe.

For wiring details see page 24.

1: Platinum

2: Not used

3: Reference

4: Shield/GND

5: Not used

6: RTD ret

7: Not used

8: RTD

9: Not used

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.4.4.1 Example 4

ORP measurement with pH Solution ground electrode (e.g. InPro 3250 SG, InPro 4800 SG).

32

7

6

5

9

8

4

3

2

Jumper

1

Not connected

Cable

h h

Temperature probe

Combination pH electrode with RTD and SG

NOTE: Jumper Terminal 3 and 4.

NOTE: Pt100 adapter required (enclosed) for sensors with Pt100 temperature probe.

For wiring details see page 24.

1: Platinum

2: Not used

3: Reference

4: Shield/GND

5: Not used

6: RTD ret

7: Not used

8: RTD

9: Not used

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.5 Connection of Analogue Sensor for

Dissolved Oxygen/Dissolved Ozone

(except 58 037 221)

4.5.1 Connecting the Sensor to the VP Cable

33

A

B

C

F

D

E

A B C D E F S h NOTE: Be sure to observe the sensor instruction manual.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.5.2 Typical Wiring using TB3/TB4

9

8

7

6

5

4

3

2

1

34 h NOTE: Wire colors only valid for connection with VP cable, blue not connected.

M300 connector:

1: not used

2: Anode

3: not used

4: Shield/GND

5: Cathode

6: NTC ret, Guard

7: Not used

8: NTC 2

9: not used

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 35

4.6 Connection of Analogue Sensor for

Dissolved Oxygen 58 037 221

This sensor consists of a Thornton Long Life dissolved oxygen probe that comes directly wired to a preamplifier box. The preamplifier connects to the M300 using a 58 080 25X series cable.

Use the connections shown in the last table of section 4.3 and follow the additional instructions supplied with the sensor.

4.7 Connection of ISM Sensor

InPro 3250i/SG/120

Art.-Nr. 52 005 373

4.7.1 Connection of ISM Sensor for pH, Cond 4-e and

Dissolved Oxygen

A B h NOTE: Connect the Sensor and screw the plug head clockwise (hand tight).

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

4.7.2 AK9 Cable Assignment

A: 1-wire data (transparent)

B: Ground/shield

4.7.3 Connection of ISM Sensor for Cond 2-e

(Thornton Models only)

36

D

A

B

C

4.7.4 Cable Assignment of ISM Sensor for Cond 2-e

(Thornton Models only)

A: GND (white)

B: Data RS485-B (black)

C: Data RS485-A (red)

D: 5 VDC (blue)

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

5 Placing transmitter in, or out, of service

5.1 Placing transmitter in service

a

After connecting the transmitter to power supply circuit, it will be active as soon as the circuit is powered.

5.2 Placing transmitter out of service

First disconnect the unit from the main power source, then disconnect all remaining electrical connections. Remove the unit from the wall/panel. Use the installation instruction in this manual as reference for dis-assembling mounting hardware.

37

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 h h h

38

6 Quick Setup

(PATH: Menu/Quick Setup)

Select Quick Setup and press the [ENTER] key. Enter the security code if necessary

(see section 9.3 “Passwords”).

NOTE: Please find the complete description of the Quick Setup routine described in the separate booklet ”Quick Setup Guide for Transmitter M300” enclosed in the box.

NOTE: Please do not use Quick Setup menu after configuration of the transmitter, because some of the parameters i.e. analog output configuration will may be reseted.

NOTE: Refer to section 3.3 “Control/Navigation Keys” for information on menu navigation.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 h

39

7

(PATH: Cal)

Sensor Calibration

The calibration key c allows the user one-touch access to Sensor calibration and verification features. Thornton models also allow access to Meter and Analog Output calibration (see chapter 11.3.1 and 11.3.2). All other models also allow access to Analog Output and Meter calibration if the access has been previously unlocked (see chapter 11.3.3 ”Calibrate Unlock”).

NOTE: During Calibration, a flashing ”H” in the upper left corner of the display indicates a calibration is in process with a Hold condition active. (The hold output function need to be activated.) See also chapter 3.3 “Display”

7.1 Enter Calibration Mode

While in Measurement mode press the security code, press the m or .

c key. If the display prompts you to enter the calibration

key to set the calibration security mode, then press the [EN-

TER] key to confirm the calibration security code.

For dual-channel devices: Using the the channel to be calibrated. Then use the m or .

key on the ”Channel A” field let you change the c key to move to the calibration field.

Press the are: m or .

key to select the type of calibration desired. The choices for each sensor type

Conductivity = Conductivity, Resistivity, Temperature*, Edit*, Verify

Oxygen** = Oxygen, Temperature*, Edit*, Verify

Ozone** = Ozone, Temperature*, Edit*, Verify pH = pH, mV, Temperature*, Edit pH*, Edit mV, Verify, ORP***

Press [ENTER].

* not on ISM version

** for Thornton transmitters (p/n 58 001 316 and 58 001 306), a jumper is required between

terminal 1, 3 and 4 on TB3 and/or TB4.

*** only with ISM sensors

After every successful calibration, the following options are available:

Calibrate: Calibration values will be overtaken und used for the measurement. Additionally, the data will be stored in the sensor*.

Abort: Calibration values will be discarded.

* only available with ISM sensors

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 h h

40

7.2 Conductivity/Resistivity Calibration

This feature provides the ability to perform a one-point, two-point or process Conductivity resp.

Resistivity ”Sensor” calibration for two- or four-electrode sensors. The procedure described below works for both types of calibrations. There is no reason to perform a two-point calibration on a

2-electrode conductivity sensor. It is also not practical to calibrate conducitivity sensors using

(low conductivity) reference solutions. It is recommended that conducitivity sensors be sent back to the factory for calibration. Consult factory for assistance.

NOTE: When performing calibration on a conductivity sensor, results will vary depending on the methods, calibration apparatus and/or quality of reference standards used to perform the calibration.

NOTE: For measuring tasks the temperature compensation for the application as defined at the menu Resitivity (resp. for M300 dual channel the menu Comp/pH/O2) will be considered and not the temperature compensation selected thru the calibration procedure (see also chapter

8.2.4.1 “Conductivity/Resistivity/ Temperature Compensation“; PATH: Menu/Configure/Measurement/Resitivity).

Enter Conductivity Sensor Calibration mode as described in section 7.1 “Enter Calibration Mode”.

When configuring Thornton model transmitters, after selecting the desired sensor calibration and pressing [ENTER], the next screen will ask to select the type of temperature compensation mode desired during the calibration process. The choices are ”None”, ”Standard”, ”Light 84”,

”Std 75 ºC”, ”Lin 20°C = 02.0%/ºC” (user selectable value), ”Lin 25°C = 02.0%/ºC” (user selectable value), ”Glycol.5”, ”Glycol1”, ”Alcohol” and “Nat H2O”.

All other models provide the options ”Standard”, ”Lin 20°C = 02.0%/ºC” (user selectable value), and ”Lin 25°C = 02.0%/ºC” (user selectable value) as Compensation mode thru the calibration.

Standard compensation includes compensation for non-linear high purity effects as well as conventional neutral salt impurities and conforms to ASTM standards

D1125 and D5391.

Lin 25°C compensation adjusts the reading by a factor expressed as ”% per °C” deviation from 25 °C. The factor can be modified.

Lin 20°C compensation adjusts the reading by a factor expressed as ”% per °C” deviation from 20 °C. The factor can be modified.

Choose the compensation mode, modify the factor where appropriate and press [ENTER]

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 41

7.2.1 One-point Sensor Calibration

(Display reflects typical Sensor calibration)

Enter Conductivity Sensor Calibration mode as described in section 7.1 “Enter Calibration Mode” and choose one of the compensation modes (see section 7.2 “Conductivity/Resistivity Calibration”).

Select 1 point Calibration and press [ENTER]. With conductitivty sensors a one-point calibration is always performed as a slope calibraton.

Place the electrode into the reference solution.

Enter the Value of calibration Point 1 and then press the [ENTER] key to start calibration. The value in the 2nd text line is the actual measured value from the sensor prior to calibration.

After the calibration the Multiplier or slope calibration factor ”M” and the Adder or offset calibration factor ”A” are displayed.

Select Yes to save the calibration values and the Successful Calibration is confirmed on the display. If an ISM sensor is connected, the calibration data will be stored in the sensor.

You will get the message ”Re-install sensor” and ”Press Enter” on the display. After pressing

”Enter” the M300 returns to the measuring mode.

7.2.2 Two-point Sensor Calibration

(4-electrode sensors only)

Enter Conductivity Sensor Calibration mode as described in section 7.1 Enter Calibration Mode and choose one of the compensation modes (see section 7.2 “Conductivity/Resistivity Calibration”).

Select 2 point Calibration and press [ENTER].

Place the electrode into the first reference solution.

CAUTION: Rinse sensors with a high-purity water solution between calibration points to prevent contamination of the reference solutions.

Enter the Value of Point 1 and press the [ENTER] key. Place the sensor into the second reference solution.

Enter the Value of Point 2 and press the [ENTER] key to start the calibration.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 42

After the calibration the Multiplier or slope calibration factor ”M” and the Adder or offset calibration factor ”A” are displayed.

Select Yes to save the calibration values and the Successful Calibration is confirmed on the display. If an ISM sensor is connected, the calibration data will be stored in the sensor.

You will get the message ”Re-install sensor” and ”Press Enter” on the display. After pressing

”Enter” the M300 returns to the measuring mode.

7.2.3 Process Calibration

Enter Conductivity Sensor Calibration mode as described in section 7.1 “Enter Calibration Mode” and choose one of the compensation modes (see section 7.2 “Conductivity/Resistivity Calibration”).

Select Process Calibration and press [ENTER]. With conductitivty sensors a process calibration is always performed as a slope calibraton.

Take a sample and press the [ENTER] key again to store the current measuring value.

During the ongoing calibration process, the letter of the channel, which is concerned by the calibration, ”A” or “B” is blinking in the display.

After determining the conductivity value of the sample, press the [CAL] key again to proceed with the calibration.

Enter the conductivity value of the sample, then press the [ENTER] key to start the calculation of calibration results.

After the calibration the Multiplier or slope calibration factor ”M” and the Adder or offset calibration factor ”A” are displayed.

Select Yes to save the calibration values and the Successful Calibration is confirmed on the display.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

7.3 Oxygen Calibration

Dissolved Oxygen calibration is performed as either a one-point or process calibration.

43

7.3.1 One-Point Sensor Calibration

Before air calibration, for highest accuracy, enter the barometric pressure and relative humidity as in section 8.2.4.3 “Dissolved Oxygen Parameters” described

Enter Oxygen Calibration mode as described in section 7.1 “Enter Calibration Mode”.

A DO sensor calibration is always either a one point Air (Slope) or a Zero (Offset) calibration. A one point slope calibration is done in air and a one point offset calibration is done at 0 ppb DO.

A one-point zero dissolved oxygen calibration is available but not normally recommended since zero DO is very hard to achieve.

Select 1 point followed by either Slope or ZeroPt as the calibration type. Press [ENTER].

h

Place the sensor in the calibration gas (e.g. air) resp. solution. Press [ENTER].

7.3.1.1 Auto mode

NOTE: For a zero point calibration the Auto mode is not available. If Auto mode has been configured (see section 8.2.4.3 “Dissolved Oxygen Parameters”) and an offset calibration will be executed, the transmitter will perform the calibration in Manual mode.

Enter the value for Point 1 including a decimal point and units. The value in the second text line is the value being measured by the transmitter and sensor in the units selected by the user.

As soon as the stabilization criteria have been fulfilled the display changes. The display shows the calibration result for slope ”S” and offset value ”Z”.

Select Yes to save the calibration values and the successful Calibration is confirmed on the display. If an ISM sensor is connected, the calibration data will be stored in the sensor.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 44

7.3.1.2 Manual mode

Enter the value for Point 1 including a decimal point and units. The value in the second text line is the value being measured by the transmitter and sensor in the units selected by the user.

Press [ENTER] when this value is stable to perform the calibration.

After the calibration the slope calibration factor S and the offset calibration factor Z are displayed.

Select Yes to save the calibration values and the successful Calibration is confirmed on the display. If an ISM sensor is connected, the calibration data will be stored in the sensor.

You will get the message ”Re-install sensor” and ”Press Enter” on the display. After pressing

”Enter” the M300 returns to the measuring mode.

7.3.2 Process Calibration

Enter Oxygen Calibration mode as described in section 7.1 “Enter Calibration Mode”.

Select Process followed by either Slope or ZeroPt as the calibration type. Press [ENTER]

Take a sample and press the [ENTER] key again to store the current measuring value. To show the ongoing Calibration Process, A or B (depending on the channel) is displayed in the top left hand corner.

After determining the O

2 bration.

Value of the Sample press the [CAL] key again to proceed with the cali-

Enter the O

2

value of the sample then press the [ENTER] key to start calibration.

After the calibration the slope calibration factor S and the offset calibration factor Z are displayed.

Select Yes to save the new calibration values and the successful Calibration is confirmed on the display. If an ISM sensor is connected, the calibration data will be stored in the sensor. The A resp. B in the top left hand corner disappears after 20 seconds.

7.4 Ozone Calibration (Thornton Models only)

Dissolved Ozone calibration is performed as a 1 point calibration and must be performed quickly because ozone decays rapidly into oxygen, especially at warm temperatures.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 45

7.4.1 One-Point Sensor Calibration

Enter Ozone Calibration mode as described in section 7.1 “Enter Calibration Mode” and select

Ozone.

Ozone sensor calibration is always either a one point Comparison (Slope) or a Zero (Offset) calibration. A one point slope calibration is always obtained from a comparison instrument or colorimetric test kit and a one point offset calibration is done in air or in Ozone-free water.

Select 1 point followed by either Slope or ZeroPt as the calibration type. Press [ENTER].

Enter the value for Point 1 including a decimal point and units. The value in the second text line is the value being measured by the transmitter and sensor in the units selected by the user.

Press [ENTER] when this value is stable to perform the calibration.

After the calibration the slope calibration factor S and the offset calibration factor Z are displayed.

h

Select Yes to save the calibration values and the successful Calibration is confirmed on the display.

You will get the message ”Re-install sensor” and ”Press Enter” on the display. After pressing

”Enter” the M300 returns to the measuring mode.

7.5 pH Calibration

For pH sensors, the M300 transmitter features one-point, two-point (Auto or Manual mode) or process calibration with 9 preset buffer sets or manual buffer entry. Buffer values refer to

25 °C. To calibrate the instrument with automatic buffer recognition, you need a standard pH buffer solution that matches one of these values (see section 8.2.4.2 ”pH Parameters” for configuring modes and selecting buffer sets). Please select the correct buffer table before using automatic calibration (see chapter 20 “Buffer tables”).

NOTE: Only buffer table 20.2.1 Mettler-pH/pNa is available for dual membrane pH electrodes

(pH/pNa)

Enter pH Calibration mode as described in section 7.1 “Enter Calibration Mode”.

7.5.1 One Point Calibration

Select 1 point Calibration. With pH sensors a one point calibration is always performed as a offset calibraton.

Depending on the parameterized Drift control (see chapter 8.2.4.2 “pH Parameters”) one of the two following modes is active.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

7.5.1.1 Auto Mode

Place the electrode in the buffer solution and press the [ENTER] key to start the calibration.

46

The display shows the buffer the transmitter has recognized (Point 1) and the measured value.

As soon as the stabilisation criteria have been fulfilled, the display changes to show the slope calibration factor S and the offset calibration factor Z.

Select Yes to save the calibration values and the successful Calibration is confirmed on the display. If an ISM sensor is connected, the calibration data will be stored in the sensor.

You will get the message ”Re-install sensor” and ”Press Enter” on the display. After pressing

”Enter” the M300 returns to the measuring mode.

7.5.1.2 Manual Mode

Place the electrode in the buffer solution. The display shows the buffer the transmitter has recognized (Point 1) and the measured value. Press [ENTER] to proceed.

The display shows now the slope calibration factor S and the offset calibration factor Z.

Select Yes to save the calibration values and the successful Calibration is confirmed on the display. If an ISM sensor is connected, the calibration data will be stored in the sensor.

You will get the message ”Re-install sensor” and ”Press Enter” on the display. After pressing

”Enter” the M300 returns to the measuring mode.

7.5.2 Two Point Calibration

Select 2 point Calibration.

Depending on the parameterized Drift control (see chapter 8.2.4.2 “pH Parameters”) one of the two following modes is active.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

7.5.2.1 Auto Mode

Place the electrode in the buffer solution and press the [ENTER] key to start the calibration.

47

The display shows the buffer the transmitter has recognized (Point 1) and the measured value.

As soon as the stabilisation criteria have been fulfilled, the display changes and prompts you to place the electrode in the second buffer.

Place the electrode in the second buffer solution and press the [ENTER] key to go on with the calibration.

The display shows the second buffer the transmitter has recognized (Point 2) and the measured value.

As soon as the stabilisation criteria have been fulfilled, the display changes to show the slope calibration factor S and the offset calibration factor Z.

Select Yes to save the calibration values and the successful Calibration is confirmed on the display. If an ISM sensor is connected, the calibration data will be stored in the sensor.

You will get the message ”Re-install sensor” and ”Press Enter” on the display. After pressing

”Enter” the M300 returns to the measuring mode.

7.5.2.2 Manual Mode

Place the electrode in the first buffer solution. The display shows the buffer the transmitter has recognized (Point 1) and the measured value. Press [ENTER] to proceed.

Place the transmitter in the second buffer solution. The display shows the buffer the transmitter has recognized (Point 2) and the measured value. Press [ENTER] to proceed.

The display shows the slope calibration factor S and the offset calibration factor Z.

Select Yes to save the calibration values and the successful Calibration is confirmed on the display. If an ISM sensor is connected, the calibration data will be stored in the sensor

You will get the message ”Re-install sensor” and ”Press Enter” on the display. After pressing

”Enter” the M300 returns to the measuring mode.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

7.5.3 Process Calibration

Select Process Calibration. At pH sensors a process calibration is always performed as a offset calibraton.

48

Take a sample and press the [ENTER] key again to store the current measuring Value. To show the ongoing Calibration Process, A or B (depending on the channel) is displayed in the top left hand corner.

After determining the pH Value of the Sample, press the [CAL] key again to proceed with the calibration.

Enter the pH value of the sample then press the [ENTER] key to start the calculation of the calibration results.

After the calibration the slope calibration factor S and the offset calibration factor Z are displayed.

Select Yes to save the new calibration values and the successful Calibration is confirmed on the display. If an ISM sensor is connected, the calibration data will be stored in the sensor. The A resp. B in the top left hand corner disappears after 20 seconds.

7.5.4 mV Calibration (not on ISM Version)

Enter mV Calibration mode as described in section 7.1 “Enter Calibration Mode”.

Enter the value for Point 1. The offset calibration factor is calculated by using the value of Point1 instead of the measured value (line 4, mV = ....) and is displayed on the next screen.

Z is the newly calculated offset calibration factor. The slope calibration factor S is always 1 and does not enter the calculation.

Select Yes to save the new calibration values and the successful Calibration is confirmed on the display.

You will get the message ”Re-install sensor” and ”Press Enter” on the display. After pressing

”Enter” the M300 returns to the measuring mode.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

7.5.5 ORP Calibration (only on ISM Version)

In case that an pH sensor with solution ground, based on the ISM technology is connected to the transmitter, the M300 ISM give the option to make in addition an ORP calibration.

Enter ORP calibration mode as described in section 7.1 “Enter Calibration Mode”.

49

Enter Point 1. In addition the actual ORP is displayed.

Press [ENTER] to proceed.

The display shows the slope calibration factor S and the offset calibration factor Z.

Select Yes to save the new calibration values and the successful Calibration is confirmed on the display. The calibration data will be stored in the sensor.

You will get the message ”Re-install sensor” and ”Press Enter” on the display. After pressing

”Enter” the M300 returns to the measuring mode.

7.6 Sensor Temperature Calibration (not on ISM Version)

Enter Calibration Mode as described in section 7.1 “Enter Calibration Mode” and select Temperature.

7.6.1 One-Point Sensor Temperature Calibration

(not on ISM Version)

Select 1 point calibration. Slope or Offset can be selected with the 1 point calibration. Select

Slope to recalculate the Slope factor M (Multiplier) or Offset to recalculate the offset calibration factor A (Adder).

Enter the value for Point 1 and press [ENTER].

The newly calculated value – either M or A – is displayed. Select Yes to save the new calibration values and the successful Calibration is confirmed on the display.

You will get the message ”Re-install sensor” and ”Press Enter” on the display. After pressing

”Enter” the M300 returns to the measuring mode.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

7.6.2 Two-Point Sensor Temperature Calibration

(not on ISM version)

Select 2 point as calibration Type.

50 h

Enter the value for Point 1 and press [ENTER].

Enter the value for Point 2 and press [ENTER].

The newly calculated values M and A are displayed. Select Yes and press [ENTER] to save the new calibration values and the successful Calibration is confirmed on the display.

You will get the message ”Re-install sensor” and ”Press Enter” on the display. After pressing

”Enter” the M300 returns to the measuring mode.

7.7 Edit Sensor Calibration Constants

(not on ISM version)

Enter Calibration mode as described in section 7.1 “Enter Calibration Mode” and select Edit,

Edit pH or Edit mV.

All calibration constants for the selected sensor channel are displayed. Primary measurement constants (p) are displayed on Line 3. Secondary measurement (temperature) constants (s) for the sensor are displayed on Line 4.

The calibration constants can be changed in this menu.

Select Yes to save the new calibration values and the successful Calibration is confirmed on the display.

NOTE: Each time a new analogue conductivity sensor is connected to the M300 transmitter, it is necessary to enter the unique calibration constant located on the sensor label.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

7.8 Sensor Verification

Enter Calibration mode as described in section 7.1 “Enter Calibration Mode” and select Verify.

51

The measured signal of the primary and the secondary measurement in electrical units are shown. The meter calibration factors are used when calculating these values.

Use the m or .

key to toggle between Channel A and B*.

* Only on dual channel version.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

8 Configuration

(PATH: Menu/Configure)

Configure

Measurement Analog Outputs Set Points Alarm/Clean Display Hold Outputs

52

8.1 Enter Configuration Mode

While in Measurement mode, press the ure – Menu and press [ENTER]. b key. Press the m or .

key to navigate to the Config-

8.2 Measurement

(PATH: Menu/Configure/Measurement)

Enter configuration mode as described in Section 8.1 ”Enter Configuration Mode”.

Press the [ENTER] key to select this Menu. The following sub menus can now be selected:

Channel Setup, Temperature Source*, Comp/pH/O2** and Set Averaging.

* Not on ISM version

** for single channel transmitters M300 and M300ISM the term in the display does not show

Comp/pH/O2 but Resistivity or pH or O2. The term depends on the version of the transmitter

M300 resp. the ISM sensor, that has been connected to the transmitter M300 ISM.

8.2.1 Channel Setup

(PATH: Menu/Configure/Measurement/Channel Setup)

Press the [ENTER] key to select the ”Channel Setup” Menu.

h

NOTE: The selection depends on the transmitter type.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 53

Analog Sensors:

Select Sensor Type and press [ENTER].

Available Sensor Types are: pH/ORP

Cond(2)

Cond(4)

O

2

O

O

2

O

2

3

hi

(I)

(V)

= pH or ORP

= 2 electrode conductivity

= 4 electrode conductivity

= Dissolved Oxygen (ppm)

= Dissolved Oxygen (except 58037221, Thornton models only)

= Dissolved Oxygen 58037221 (Thornton models only)

= Dissolved Ozone (Thornton models only)

ISM Sensors: pH/ORP pH/pNa

O

2

hi

Cond (2)

Cond (4)

= pH or ORP

= pH and ORP (with pH/pNa electrode)

= Dissolved oxygen (ppm)

= Cond 2-e sensor (Thornton models only)

Auto:

If you select a specific parameter instead of auto, the transmitter only accepts the selected parameter type.

= Cond 4-e sensor

= The transmitter automatically recognizes the connected sensor

The 4 lines of the display can now be configured with sensor channel ”A” or ”B” for each line of the display as well as measurements and unit multipliers. Pressing the [ENTER] key will display the selection for lines c and d.

Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard the entered values and return to the measurement display screen, selecting Yes will save changes made.

h

8.2.2 Derived Measurements (Thornton models only)

There are three derived measurements available for configuration with two Conductivity sensors:

%Rej (% Rejection), pH Cal (Calculated pH) and CO

2

Cal (Calculated CO

2

). To set up any of the derived measurements, first set up the two primary conductivity measurements, which will be used to calculate the derived measurement. Define the primary measurements as if they were stand-alone readings. Then the derived measurement can be defined.

NOTE: It is important to use the same units for both measurements.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 h h

54

8.2.2.1 % Rejection measurement

For reverse osmosis (RO) applications, percent rejection is measured with conductivity to determine the ratio of impurities removed from product or permeate water to the total impurities in the incoming feed water. The formula for obtaining Percent Rejection is:

[1 – (Product/Feed)] X 100 = % Rejection

Where Product and Feed are the conductivity values measured by the respective sensors.

Figure 4.1 shows a diagram of an RO installation with sensors installed for Percent Rejection.

Conductivity

Sensor

B

Reverse Osmosis Membrane

Conductivity

Sensor

A

Feed Product

Reject

Figure 4.1: % Rejection

NOTE: The product monitoring sensor must be on the channel that will measure percent rejection. If the product conductivity sensor is installed in channel A, then percent rejection must be measured in channel A.

8.2.2.2 Calculated pH (Power plant applications only)

Calculated pH may be obtained very accurately from specific and cation conductivity values on power plant samples when the pH is between 7.5 and 10.5 due to ammonia or amines and when the specific conductivity is significantly greater than the cation conductivity. This calculation is not suitable where significant levels of phosphates are present. The M300 uses this algorithm when pH CAL is selected as a measurement.

The calculated pH must be configured on the same channel as specific conductivity. For example, set up measurement ”a” on channel A to be specific conductivity, measurement ”b” on

Channel B to be cation conductivity, measurement ”c” on channel A to be calculated pH and measurement ”d” on channel A to be temperature. Set the temperature compensation mode to

”Ammonia” for measurement ”a” and to ”Cation” for measurement ”b”.

NOTE: If operation goes outside the recommended conditions, a glass electrode pH measurement is needed to obtain an accurate value. On the other hand, when sample conditions are within the ranges noted above, the calculated pH provides an accurate standard for one-point trim calibration of the electrode pH measurement.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 h h

55

8.2.2.3 Calculated CO

2

(Power plant applications only)

Carbon dioxide may be calculated from cation conductivity and degassed cation conductivity measurements on power plant samples using tables from ASTM Standard D4519. The M300 has these tables stored in memory, which it uses when units of CO

2

CAL are selected.

The calculated CO

2

measurement must be configured to the same channel as cation conductivity. For example, set up measurement ”a” on channel A to be cation conductivity, measurement

”b” on channel B to be degassed cation conductivity, measurement ”c” on channel A to be calculated CO

2

and measurement ”d” on channel B to be temperature.

Set the temperature compensation mode to ”Cation” for both conductivity measurements.

8.2.3 Temperature Source (not on ISM version)

(PATH: Menu/Configure/Measurement/Temperature Source)

Enter Configuration Mode as described in section 8.1 ”Enter Configuration mode” and select the menu Measurement (see section 8.2 ”Configuration/Measurement”).

Navigate to the menu Temperature Source by using the select this Menu. The following options can be chosen: m

”Fixed”: allows a specific temperature value to be entered.

or .

key. Press the [ENTER] key to

NOTE: The selection depends on the transmitter type. For detailed overview see the specifications in section 16 ”Specification”.

”Use this Ch Pt1000”: temperature input will be taken from the sensor attached.

”Use this Ch Pt100”: temperature input will be taken from sensor attached

”Use this NTC22K”: temperature will be taken from the sensor attached

”Fixed = 25°C”: allows a specific temperature value to be entered

”Use Other Channel”: temperature input will be taken from the sensor attached to the other channel (only on dual channel version)

NOTE: If temperature source is set to Fixed, the temperature applied during one-point and/or twopoint calibration of pH electrodes can be adjusted within the corresponding calibration procedure. After the calibration the fixed temperature defined in this configuration menu is valid again.

Pressing the [ENTER] key will bring up the Save Changes dialog.

Selecting No will discard the entered values and return to the measurement display screen, selecting Yes will save changes made.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 h

56

8.2.4 Parameter Related Settings

(PATH: Menu/Configure/Measurement/Comp/pH/O2)

Additional measurement and calibration parameters can be set for each parameter; conductivity, pH and O2.

NOTE: Use pH menu for settings of pH/pNa sensors.

Enter Configuration Mode as described in section 8.1 ”Enter Configuration mode” and select the menu Measurement (see section 8.2 ”Configuration/Measurement”).

For dual channel devices: The menu Comp/pH/O2 can be selected by using the

Then use the c m or

key to move to the next input field and select the parameter using the

.

m

key.

or key. The choices are Resistivity (for conductivity measurement), pH and O2. Press [ENTER]

.

For single-channel devices: The menu can be selected by using the m or play: Resistivity (for conductivity measurement), pH or O2. Press [ENTER]

.

key. Depending on the connected ISM sensor resp. used transmitter M300 the following term is shown in the dis-

For more details, please see the following explanations according to the different parameters.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 h

8.2.4.1 Conductivity/Resistivity Temperature Compensation

57

NOTE: The full Temperature Compensation selection is available on Thornton model transmitters only. All other models provide Standard, Lin 25°C or Lin 20°C compensation.

Select Resistivity and press [ENTER].

The temperature compensation mode for any of the four measurement lines can be selected.

Temperature compensation should be matched to the characteristics of the application.

Choices are ”None”*, ”Standard”, ”Light 84”*, ”Std 75 ºC”*, ”Lin 25°C”, ”Glycol.5”*, ”Glycol1”*,

”Cation”*, ”Alcohol”*, ”Ammonia”* and ”Lin 20°C”.

Standard compensation includes compensation for non-linear high purity effects as well as conventional neutral salt impurities and conforms to ASTM standards D1125 and D5391.

* Std 75 °C compensation is the Standard compensation algorithm referenced to 75°C. This compensation may be preferred when measuring Ultrapure Water at an elevated temperature.

(Resistivity of ultrapure water compensated to 75 °C is 2.4818 Mohm-cm.

Lin 25 °C compensation adjusts the reading by a factor expressed as a ”% per °C” (deviation from 25 °C). Use only if the sample has a well-characterized linear temperature coefficient. The factory default setting is 2.0%/°C.

* Glycol.5 compensation matches the temperature characteristics of 50% ethylene glycol in water. Compensated measurements using this solution may go above 18 Mohm-cm.

* Glycol1 compensation matches the temperature characteristics of 100% ethylene glycol.

Compensated measurements may go well above 18 Mohm-cm.

* Cation compensation is used in power industry applications measuring the sample after a cation exchanger. It takes into account the effects of temperature on the dissociation of pure water in the presence of acids.

* Alcohol compensation provides for the temperature characteristics of a 75% solution of isopropyl alcohol in pure water. Compensated measurements using this solution may go above 18 Mohm-cm.

* Light 84 compensation matches the high purity water research results of Dr. T.S. Light published in 1984. Use only if your institution has standardized on that work.

* Ammonia compensation is used in power industry applications for specific conductivity measured on samples using ammonia and/or ETA (ethanolamine) water treatment. It takes into account the effects of temperature on the dissociation of pure water in the presence of these bases.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 58

Lin 20 °C compensation adjusts the reading by a factor expressed as a ”% per °C” (deviation from 20 °C). Use only if the solution has a well-characterized linear temperature coefficient. The factory default setting is 2.0% / °C.

If compensation mode ”Lin 25 °C” or ”Lin 20 °C” has been chosen, the factor for the adjustment of the reading can be modified after pressing [ENTER] (If working at measurement line a or b press [ENTER] twice).

Pressing [ENTER] will bring up the Save Changes dialog. Selecting No will discard the entered values and return to the measurement display screen, selecting Yes will save changes made.

* On Thornton models only.

8.2.4.2 pH/ORP Parameters

Select pH and press [ENTER].

h

Select the drift control for calibration as Auto (stabilisation and time criteria have to be fulfilled) or Manual (the user can decide when a signal is stable enough to complete calibration) followed by the relevant buffer table for the automatic buffer recognition. If the drift rate is less than 0.8 mV over a 20 second interval then the reading is stable and the calibration is done using the last reading. If the drift criteria is not met within 300 seconds then the calibration times out and the message ”Calibration not done” is displayed. Press [ENTER].

For automatic buffer recognition during calibration, select the buffer solution set that will be used:

Mettler-9, Mettler-10, NIST Tech, NIST Std, HACH, CIBA, MERCK, WTW, JIS Z 8802 or None. See

Section 20 ”Buffer Tables” for buffer values. If the auto buffer feature will not be used or if the available buffers are different from those above, select None. Press [ENTER].

NOTE: For dual membrane pH electrodes (pH/pNa) only buffer Na+ 3.9M (see section 19.2.1

“Mettler-pH/pNa buffers”) is available.

STC is the solution temperature coefficient in units of pH/°C referenced to 25 °C (Default = 0.000 for most applications). For pure waters, a setting of – 0.016 pH/°C should be used. For low conductivity power plant samples near 9 pH, a setting of – 0.033 pH/°C should be used. These negative coefficients compensate for the negative temperature influence on the pH of these samples. Press [ENTER].

IP is the isothermal point value (Default = 7.000 for most applications). For specific compensation requirements or non standard inner buffer value, this value can be changed.

Press [ENTER].

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 59

STC RefTemp sets the temperature to which solution temperature compensation is referenced.

The displayed value and the output signal is referenced to STC RefTemp. Selecting ”No” means solution temperature compensation is not used. The most common reference temperature is

25°C. Press [ENTER].

The units for the slope and the zero point, that will be shown on the display can be chosen. The default setting for the unit of the slope is [%] and can be changed to [pH/mV]. For the zero point the default setting of the unit is [pH] and can be changed to [mV]. Use the to the input field and select the unit by using the m or .

key. c key to move

Pressing [ENTER] again will bring up the Save Changes dialog. Selecting No will discard the entered values and return to the measurement display screen, selecting Yes will save changes made.

8.2.4.3 Dissolved Oxygen Parameters

Select O

2

and press [ENTER]

Enter the Calibration pressure. The default value for CalPres is 759.8 and the default unit is mmHg. Press [ENTER]

Enter the Process Pressure. The units for ProcPres and CalPres do not have to be the same.

Press [ENTER]

For the algorithm of the process calibration the applied pressure (ProcCalPres) has to be defined. The value of the process pressure (ProcPres) or the calibration pressure (CalPres) can be used. Chose the pressure, that applies during the process calibration, resp. should be used for the algorithm and press [ENTER]

Select the required Drift Control of the measuring signal during the calibration procedure. Choose

Manual if the user will decide when a signal is stable enough to complete the calibration. Select

Auto and an automatic stability control of the sensor signal during calibration through the transmitter will be done. Press [ENTER]

In the next step the salinity of the measured solution can be modified. Press [ENTER]

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

In addition the relative humidity of the calibration gas can also be entered. The allowed values for relative humidity are in the range 0% to 100%.

Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard the entered values and return to the measurement display screen, selecting Yes will save changes made.

60

8.2.5 Set Averaging

(PATH: Menu/Configure/Measurement/Set Averaging)

Press the [ENTER] key to select this Menu. The averaging method (noise filter) for each measurement line can now be selected. The options are Special (Default), None, Low, Medium and

High:

None = no averaging or filtering

Low = equivalent to a 3 point moving average

Medium = equivalent to a 6 point moving average

High = equivalent to a 10 point moving average

Special = averaging depending on signal change (normally High averaging but Low averaging for large changes in input signal)

Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard the entered values and return to the measurement display screen, selecting Yes will save changes made.

8.3 Analog Outputs

(PATH: Menu/Configure/Analog Outputs)

Enter configuration mode as described in Section 8.1 ”Enter Configuration Mode” and navigate to the menu “Analog Outputs” by using the m or .

key.

Press the [ENTER] key to select this Menu, which lets you configure the 2 (4 for dual channel version) Analog Outputs.

Once analog outputs have been selected, use the ble: b and c buttons to navigate between configurable parameters. Once a parameter is selected, its setting can be selected per the following ta-

When an Alarm Value is selected, the analog output will go to this value if any alarm condition occurs.

Parameter Selectable Values

Aout: 1, 2, 3* or 4* (default is 1)

Measurement: a, b, c, d or blank (none) (default is blank)

Alarm Value: 3.6 mA, 22.0 mA or Off (default is off)

* Only on dual channel version.

The Aout type can be Normal, Bi-Linear, Auto-Range or Logarithmic. The range can be 4–20mA or 0–20mA. Normal provides linear scaling between the minimum and maximum scaling l

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 imits and is the default setting. Bi-Linear will also prompt for a scaling value for the mid-point of the signal and allows two different linear segments between the minimum and maximum scaling limits.

Enter the minimum and maximum Value of Aout.

61

If Auto-range was selected then Aout max1 can be configured. Aout max1 is the maximum value for the first range on Auto-Range. The maximum value for the second range on Auto-

Range was set in the previous menu. If Logarithmic Range was selected, it will also prompt for the number of decades as ”Aout1 # of Decades =2”.

The value for the Hold mode can be configured to hold the Last value or can be set to a Fixed value.

Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard the entered values and return to the measurement display screen, selecting Yes will save changes made.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

8.4 Setpoints

(PATH: Menu/Configure/Setpoints)

Enter configuration mode as described in Section 8.1 ”Enter Configuration Mode”.

Press the [ENTER] key to select this Menu.

62

4 (6 for dual channel version) Setpoints can be configured on any of the measurements (a thru d). The possible Setpoint types are Off, High, Low, Outside and Between. Thornton models also include types, %USP, %EP PW and %EP WFI for configuration with conductivity sensors.

An ”Outside” Setpoint will cause an alarm condition whenever the measurement goes above its high limit or below its low limit. A ”Between” Setpoint will cause an alarm condition to occur whenever the measurement is between its high and low limits.

USP and EP Setpoints on Thornton models provide a high alarm used for pharmaceutical water monitoring with non-temperature compensated conductivity measurements. USP (United States

Pharmacopoeia) section < 645 > and European Pharmacopoeia require that non-temperature compensated conductivity of pharmaceutical waters must be below a limit from tables based on the temperature of the sample. In other words, pharmaceutical requirements temperaturecompensate the limit rather than the measurement.

The Mettler Toledo Thornton M300 has these pharmaceutical limit tables in memory and automatically determines the conductivity limit based on the measured temperature. USP and EPWFI

(Water for Injection) setpoints use Table 8.1. The limit is the conductivity value corresponding to the 5 °C temperature step immediately below or equal to the measured temperature value. EP

Highly Purified Water limits are identical to EP WFI limits.

EP PW (Purified Water) setpoints use Table 8.2. The limit in this case is the conductivity value interpolated for the measured temperature. The M300 takes care of this automatically.

The pharmaceutical setpoint value entered into the M300 is the percentage safety margin below the limits to activate the setpoint. For example, the USP table conductivity limit at 15 °C is 1.0 µS/cm. If the setpoint value is set at 40% then the setpoint will activate whenever the conductivity goes above 0.6 µS/cm at 15 °C.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 63

Table 8.1: USP Section < 645 > Stage 1, EP WFI (Water for Injection), and EP Highly Purified Water Conductivity Limits as a Function of Temperature.

70

75

80

85

90

95

100

35

40

45

50

55

60

65

Temperature (ºC) Conductivity Limit (µS/cm)

0 0.6

5

10

0.8

0.9

15

20

25

30

1.0

1.1

1.3

1.4

1.5

1.7

1.8

1.9

2.1

2.2

2.4

2.5

2.7

2.7

2.7

2.7

2.9

3.1

Table 8.2: EP PW (Purified Water) Conductivity Limits as a Function of Temperature

40

50

60

70

75

80

90

100

Temperature (ºC) Conductivity Limit (µS/cm)

0 2.4

10

20

25

30

3.6

4.3

5.1

5.4

6.5

7.1

8.1

9.1

9.7

9.7

9.7

10.2

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 64

Enter the desired value(s) for the Setpoint and press [ENTER]

This screen provides the option to configure a setpoint to be active on an over range condition.

Select the setpoint and ”Yes” or ”No”. Select the desired relay that will activate when the setpoint alarm condition is reached.

Out of Range

Once configured, the selected relay will be activated if a sensor Out of Range condition is detected on the assigned input channel.

Delay

Enter the delay time in seconds. A time delay requires the setpoint to be exceeded continuously for the specified length of time before activating the relay. If the condition disappears before the delay period is over, the relay will not be activated.

Hysteresis

Enter value for the hysteresis. A hysteresis value requires the measurement to return within the setpoint value by a specified hysteresis before the relay is deactivated.

For a high setpoint, the measurement must decrease more than the indicated hysteresis below the setpoint value before the relay is deactivated. With a low setpoint, the measurement must rise at least the hysteresis above the setpoint value before the relay is deactivated. For example, with a high setpoint of 100 and hysteresis of 10, when this value is exceeded, the measurement must fall below 90 before the relay is deactivated.

Hold

Enter the Relay Hold Status of ”Last”, ”On” or ”Off”. This is the state the Relay will go to during a

Hold status.

State

Relay contacts are in normal state until the associated setpoint is exceeded, then the relay is activated and the contact states change.

Select ”Inverted” to reverse the normal operating state of the relay (i.e. Normally open contacts are in a closed state, and normally closed contacts are in an open state, until the setpoint is exceeded). ”Inverted” relay operation is functional when power is applied to the M300 transmitter.

Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard the entered values and return to the measurement display screen, selecting Yes will save changes made.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

8.5 Alarm/Clean

(PATH: Menu/Configure/Alarm/Clean)

Enter configuration mode as described in Section 8.1 ”Enter Configuration Mode”.

This Menu allows the configuration of Alarm and Clean functionality.

65

8.5.1 Alarm

To select ”Setup Alarm”, press the m or .

key so that ”Alarm” is flashing.

Using the b and c buttons, navigate to ”Use Relay #”. Using the to be used for the Alarm and press [ENTER].

m or .

keys, select a relay

One of the following events may be alarmed:

1. Power Failure

2. Software Failure

3. Rg Diagnostics – pH glass membrane resistance (pH sensors only; pH/pNa Rg diagnostics detect both pH and pNa membrane glasses)

4. Rr Diagnostics – pH reference resistance (pH sensors only; except pH/pNa)

5. Cond cell open (cond sensors only)

6. Cond cell shorted (cond sensors only)

7. Channel A disconnected (ISM sensors only)

8. Channel B disconnected (only for ISM sensors and dual channel version)

If any of these criterias are set to Yes and the conditions for an alarm are given, the flashing symbol a will be shown in the display, an alarm message will be recorded (see also chapter

12.1 ”Messages”; PATH: Info/Messages) and the selected relay will be activated. Furthermore an alarm can be indicated by the current output if this has been parameterized (see chapter 8.3

”Analog Outputs”; PATH: Menu/Configure/Analog Outputs)

1. there is a power failure or power cycling

2. the software watchdog performs a reset

3. Rg is out of tolerance – for example, broken measuring electrode (pH only; pH/pNa Rg diagnostics detect both pH and pNa membrane glasses)

4. Rr is out of tolerance – for example, coated or depleted reference electrode (pH only)

5. If the conductivity sensor is on air (for example in an empty pipe)

6. If the conductivity sensor has a short cut

7. If no sensor is connected on channel A (ISM sensors only)

8. If no sensor is connected on channel B (only for ISM sensors and dual channel version)

For 1 and 2 the alarm indicator will be turned off when the alarm message is cleared. It will reappear if the power is constantly cycling or if the watchdog is repeatedly resetting the system.

Only for pH sensors

For 3 and 4 the alarm indicator will go off if the message is cleared and the sensor has been replaced or repaired so that the Rg and Rr values are within specification. If the Rg or Rr message is cleared and Rg or Rr is still out of tolerance then the alarm will stay on and the message will reappear. The Rg and Rr alarm can be turned off by going into this menu and setting Rg Diagnostics and/or Rr Diagnostics to No. The message can then be cleared and the alarm indicator will be off even though Rg or Rr is out of tolerance.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 h

66

Each Alarm Relay can be configured in either a Normal or Inverted state. Select ”Inverted” to reverse the normal operating state of the relay (i.e. Normally open contacts are in a closed state, and normally closed contacts are in an open state, until an alarm occurs). ”Inverted” relay operation is functional when power is applied to the M300 transmitter.

In addition, a Delay for the activation can be set. Enter the delay time in seconds. A time delay requires the alarm to occur continuously for the specified length of time before activating the relay. If the alarm disappears before the delay period is over, the relay will not be activated.

If power failure is turned on, only inverted state is possible and cannot be changed.

Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard the entered values, selecting Yes will make the entered values the current ones.

Please note, that there are additional alarms, which will be indicated in the display. See therefore in chapter 14 ”Troubleshooting” the different warning- and alarm lists.

8.5.2 Clean

Configure the Relay to be used for the cleaning cycle.

The Default value is Relay 1.

The Cleaning Interval can be set from 0.000 to 999.9 hours. Setting it to 0 turns the clean cycle off. The cleaning time can be 0 to 9999 seconds and must be smaller than the Cleaning

Interval.

Select the desired Relay state: Normal or Inverted.

Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard the entered values and return to the measurement display screen, selecting Yes will save changes made.

8.6 Display

(PATH: Menu/Configure/Display)

Enter configuration mode as described in Section 8.1 ”Enter Configuration Mode”.

This Menu allows for the configuration of the values to be displayed and also the configuration of the Display itself.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

8.6.1 Measurement

The Display has 4 lines. Line 1 on top and Line 4 on the bottom.

Select the values (Measurement a, b, c or d) to be displayed on each line of the display.

The selection of the values for a, b, c, d needs to be done under Configuration/Measurement/

Channel Setup.

67

Select the ”Error Display” mode. If this is set to ”On” when an alarm has occurred, the message

”Failure – Press Enter” will be displayed on Line 4 when an alarm occurs in the normal Measurement mode.

Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard the entered values, selecting Yes will make the entered values the current ones.

8.6.2 Resolution

This menu allows the setting of the resolution of each displayed value.

Possible settings are 1, 0.1, 0.01, 0.001 or Auto.

Pressing the [ENTER] key will bring up the Save Changes dialog.

8.6.3 Backlight

This Menu allows the setting of the back light options of the display.

Possible settings are On, On 50% or Auto Off 50%. If Auto Off 50% is selected then the backlight will go to 50% after 4 minutes with no keypad activity. The backlight will automatically come back on if a key is pressed.

Pressing the [ENTER] key Will bring up the Save Changes dialog.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 68

8.6.4 Name

This menu allows for the configuration of an alpha-numeric name which is displayed in the first

9 characters on Lines 3 and 4 of the Display. The default is nothing (blank).

If a name is entered on line 3 and/or 4 a measurement can be still displayed on the same line.

Use the b and c keys to navigate between digits to be altered. Using the tered, press [ENTER] to bring up the Save Changes dialog.

m and .

keys to change the character to be displayed. Once all digits of both display channels have been en-

The resulting display in the measurement mode appears on Lines 3 and 4 ahead of the measurements.

h

8.7 Hold Analog Outputs

(PATH: Menu/Configure/Hold Outputs)

Enter configuration mode as described in Section 8.1 ”Enter Configuration Mode”.

The ”Hold outputs” function applies during the calibration process. If set ”Hold outputs” to Yes, during calibration process the analog output, the output relay and USB ouptut will be at hold state. The hold state depends on the setting. For the possible hold settings, see the list below.

The following options are possible:

Hold Outputs? Yes/No

The ”DigitalIn” function applies all the time. As soon as a signal is active on the digital input the transmitter goes to hold mode and the values on the analog output, the output relays and the

USB output will be at hold state.

DigitalIn1/2* State = Off/Low/High

NOTE: DigitalIn1 is to hold channel A

DigitalIn2 is to hold channel B*

* Only on dual channel version.

Possible Hold states:

Output relays:

Analog Output:

USB:

PID relay:

PID analog:

On/Off

Last/Fixed

Last/Off

Last/Off

Last/Off

(Configuration/Set point)

(Configuration/Analog output)

(System/USB)

(PID setup/Mode)

(PID setup/Mode)

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

9 System

(PATH: Menu/System)

Set Language USB

System

Passwords

Set/Clear

Lockout

Reset

69

While in Measurement mode press the

Menu and press [ENTER].

b key. Press the .

or m key to navigate to ”System” –

9.1 Set Language

(PATH: Menu/System/Set Language)

This Menu allows the configuration of the Display language.

The following selections are possible:

English, French, German, Italian, Spanish, Russian, Portuguese and Japanese.

Pressing the [ENTER] key will bring up the Save Changes dialog.

9.2 USB

(PATH: Menu/System/USB)

This menu allows configuration of the USB hold function.

USB Hold may be set to either Off or Last Values. An external host device may poll the M300 for data. If the USB Hold is set to Off, current values are returned. If the USB Hold is set to Last

Values, the values present at the time the hold condition was established are returned.

Press [ENTER] to bring up the Save Changes dialog.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 70

9.3 Passwords

(PATH: Menu/System/Passwords)

This Menu allows for the configuration of Operator and Administrator Passwords, as well as setting up a List of allowed Menus for the Operator. The Administrator has rights to access all

Menus. All default passwords for new transmitters are ”00000”.

The Passwords Menu is protected: Enter the Administrator Password to enter the Menu.

9.3.1 Changing Passwords

See Section 9.3 ”Passwords” how to enter the Passwords Menu. Select Change Administrator or

Change Operator and set the new Password.

Press the [ENTER] key and confirm the new password. Press [ENTER] again to bring up the

Save Changed dialog.

9.3.2 Configuring Menu Access for Operator

See 9.3 ”Passwords” how to enter the Passwords Menu. Select Configure Operator to configure the Access list for the Operator. It is possible to assign/deny rights to the following Menus: Cal

Key, Quick Setup, Configuration, System, PID Setup and Service.

Choose either Yes or No to give/deny access to the above Menus and press [ENTER] to advance to the next items. Pressing the [ENTER] key after configuring all menus will bring up the Save

Changes dialog. Selecting No will discard the entered values, selecting Yes will make the entered values the current ones.

9.4 Set/Clear Lockout

(PATH: Menu/System/Set/Clear Lockout)

This menu enables/disables the Lockout functionality of the transmitter. The user will be asked for a password before being allowed into any menus if the Lockout functionality is enabled.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 71

The Lockout – Menu is protected: Enter the Administrator Password and select YES to enable or

NO to disable the Lockout functionality. Pressing the [ENTER] key after the selection will bring up the Save Changes dialog. Selecting No will discard the entered value, selecting Yes will make the entered value the current one.

9.5 Reset

(PATH: Menu/System/Reset)

This Menu allows access to the following options:

Reset System, Reset Meter Cal*, Reset Analog Cal, Reset Cal Data**.

* Not on ISM version

** For ISM version of Cond 2-e sensors only.

9.5.1 Reset System

This Menu allows the reset of the meter to the factory default settings (Setpoints off, analog outputs off, etc.). The meter calibration and the analog output calibration are not affected.

Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No will return the user to the Measurement mode with no changes. Selecting Yes will reset the meter.

9.5.2 Reset Meter Calibration (not on ISM version)

This Menu allows the reset of the meter”s calibration factors to the last factory calibration values.

Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No will return the user to the Measurement mode with no changes. Selecting Yes will reset the meter calibration factors.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 72

9.5.3 Reset Analog Calibration

This Menu allows reset of the Analog Output calibration factors to the last factory calibration values.

Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No will return the user to the Measurement mode with no changes. Selecting Yes will reset the Analog Output calibration.

h

9.5.4 Reset Sensor Calibration Data to Factory Settings

If a Cond 2-e sensor based on ISM technology is connected to the transmitter, this menu is available. The menu allows the reset of the calibration data (M resp. A) of the sensors to the factory settings.

Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No will return the user to the Measurement mode with no changes. Selecting Yes will reset the calibration data of the sensor to factory settings.

NOTE: To ensure best measuring results, a new calibration of the sensor is recommended after a reset of the calibration data to factory settings. Depending on the application, the calibration could be temporarily performed as a process calibration, recommended however is an one point calibration (see chapter 7.2 “Conductivity/Resistivity Calibration”).

9.5.5 Reset Calibration Data of Sensor Electronic to Factory

Settings

If a Cond 2-e sensor based on ISM technology is connected to the transmitter, this menu is available. The menu allows the reset of the calibration data of the sensor`s evaluation electronic to the factory settings.

The procedure for this functions is described in the sensor manual.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

10 PID Setup

(PATH: Menu/PID Setup)

PID Setup

PID A/M

Tune

Parameters

Mode

73

PID Display

Setup

PID control is proportional, integral and derivative control action that can provide smooth regulation of a process. Before configuring the transmitter, the following process characteristics must be identified.

Identify the control direction of the process

– Conductivity:

Dilution – direct acting where increasing measurement produces increasing control output such as controlling the feed of low conductivity diluting water to rinse tanks, cooling towers or boilers

Concentrating – reverse acting where increasing measurement produces decreasing control output, such as controlling chemical feed to attain a desired concentration

– Dissolved Oxygen:

Deaeration – direct acting where increasing DO concentration produces increasing control output such as controlling the feed of a reducing agent to remove oxygen from boiler feedwater

Aeration – reverse acting where increasing DO concentration produces decreasing control output, such as controlling an aerator blower speed to maintain a desired DO concentration in fermentation or wastewater treatment

– pH/ORP:

Acid feed only – direct acting where increasing pH produces increasing control output, also for ORP reducing reagent feed

Base feed only – reverse acting where increasing pH produces decreasing control output, also for ORP oxidizing reagent feed

Both acid and base feed – direct and reverse acting

– Ozone:

Ozone destruct – direct acting where increasing ozone concentration produces increasing control output such as increasing UV lamp intensity

Ozonation – reverse acting where increasing ozone concentration produces a decreasing control output to decrease the output of an ozonator.

Identify the control output type based on the control device to be used:

Pulse Frequency – used with pulse input metering pump

Pulse Length – used with solenoid valve

Analog – used with current input device such as electric drive unit, analog input metering pump or current-to-pneumatic (I/P) converter for pneumatic control valve

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

120%

100%

80%

60%

40%

20%

0%

–20%

–40%

–60%

–80%

–100%

–120%

0

74

Default control settings provide linear control, which is appropriate for conductivity, dissolved oxygen and ozone. Therefore, when configuring PID for these parameters (or simple pH control) ignore settings of deadband and corner points in the Tuning Parameter section below. The nonlinear control settings are used for more difficult pH/ORP control situations.

If desired, identify the non-linearity of the pH/ORP process. Improved control can be obtained if the non-linearity is accommodated with an opposing non-linearity in the controller. A titration curve (graph of pH or ORP vs. reagent volume) made on a process sample provides the best information. There is often a very high process gain or sensitivity near the setpoint and decreasing gain further away from the setpoint. To counteract this, the instrument allows for adjustable non-linear control with settings of a deadband around the setpoint, corner points further out and proportional limits at the ends of control as shown in the figure below. Determine the appropriate settings for each of these control parameters based on the shape of the pH process titration curve.

Controller with Corner Points

Proportional limit +100% value

2

Direct corner point (value, %)

Set point value

Deadband + value

Deadband - value

Reverse corner point (value, %)

4 6

Proportional limit –100% value

8

Process Variable

10 12 14

10.1 Enter PID Setup

(PATH: Menu/PID Setup)

While in Measurement mode press the up – Menu and press [ENTER].

b key. Press the m or .

key to navigate to the PID Set-

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 75

10.2 PID Auto/Manual

(PATH: MENU/PID Setup/PID A/M)

This menu allows selection of Automatic or Manual operation. Select Auto or Manual operation.

Pressing the [ENTER] key will bring up the Save Changes dialog.

10.3 Mode

(PATH: MENU/PID Setup/Mode)

This menu contains the selection of control modes using relays or analog outputs.

Press [ENTER]. h

10.3.1 PID Mode

This menu assigns a relay or analog output for PID control action as well as details of their operation. Based on the control device being used, select one of the following three paragraphs for use with solenoid valve, pulse input metering pump or analog control.

Pulse Length – If using a solenoid valve, select ”Relays” and ”PL”, Pulse Length.

Choose the first relay position as #3 (recommended) and/or the second relay position as #4 (recom mended) as well as the Pulse Length (PL) according to the table below. A longer pulse length will reduce wear on the solenoid valve. The % ”on” time in the cycle is proportional to the control output.

NOTE: All relays from #1 to #6 can be used for the controlling function.

Conductivity

1 st Relay Position

(#3)

Controlling concentrating reagent feed pH/ORP

Dissolved

Oxygen

Dissolved

Ozone

Feeding base

Reverse control action not recommended

2 nd Relay Position

(#4)

Pulse Length (PL)

Controlling dilution water

Feeding acid

Direct acting control action

Short (PL) provides more uniform feed.

Suggested start point

= 30 sec

Reagent addition cycle: short PL provides more uniform addition of reagent. Suggested start point = 10 sec

Feed cycle time: short

PL provides more uniform feed. Suggested start point = 30 sec not recommended

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 h

76

Pulse Frequency – If using a pulse input metering pump, select ”Relays” and ”PF”, Pulse Frequency. Choose the first relay position as #3 and/or the second relay position as #4 according to the table below. Set the pulse frequency to the maximum frequency allowed for the particular pump being used, typically 60 to 100 pulses/minute. Control action will produce this frequency at 100% output.

NOTE: All relays from #1 to #6 can be used for the controlling function.

CAUTION: Setting the Pulse Frequency too high may cause the pump to overheat.

Conductivity pH/ORP

Dissolved

Oxygen

1 st Relay Position = #3 2 nd Relay Position = #4 Pulse Frequency (PF)

Controlling concentrating chemical feed

Controlling dilution water

Max allowed for the pump used (typically

60–100 pulses/minute)

Feeding base

Reverse control action

Feeding acid

Direct acting control action

Max allowed for the pump used (typically

60–100 pulses/minute)

Max allowed for the pump used (typically

60–100 pulses/minute)

Dissolved

Ozone not recommended not recommended

Analog – If using Analog control, change ”Relays” to ”Analogout” using up/down arrow keys.

Choose the first Analogout position as #1 and/or the second Analogout position as #2 according to the table below. Select the analog output current range required by the control device,

4–20 or 0–20 mA. Press [ENTER].

Conductivity pH/ORP

Dissolved Oxygen

Dissolved Ozone

1 st Analogout Position = #1

Controlling concentrating chemical feed

Feeding base

Reverse control action

Controlling ozonation

2 nd Analogout Position = #2

Controlling dilution water

Feeding acid

Direct acting control action

Controlling ozone destruction

10.4 Tune Parameters

(PATH: MENU/PID Setup/Tune Parameters)

This menu assigns control to a measurement and sets the setpoint, tuning parameters and nonlinear functions of the controller through a series of screens.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 77

10.4.1 PID Assignment & Tuning

Assign the measurement, a, b, c, or d to be controlled after ”PID on_”. Set the Gain (unitless),

Integral or Reset time Tr (minutes) and Rate or Derivative time Td (minutes) needed for control.

Press [ENTER]. Gain, Reset and Rate are later adjusted by trial and error based on process response. Always begin with Td at zero.

10.4.2 Setpoint & Deadband

Enter the desired setpoint value and the deadband around the setpoint, where no proportional control action will take place. Be sure to include the units multiplier µ or m for conductivity.

Press [ENTER].

10.4.3 Proportional Limits

Enter the low and high proportional limits – the range over which control action is required.

Be sure to include the units multiplier µ or m for conductivity. Press [ENTER].

10.4.4 Corner Points

Enter the low and high corner points in conductivity, pH, dissolved oxygen or dissolved ozone units and the respective output values from –1 to +1, shown in the figure as –100 to +100%.

Press [ENTER].

10.5 PID Display

(PATH: Menu/PID Setup/PID Display Setup)

This screen enables display of PID control status in the normal measurement mode.

When PID Display is selected, the status (Man or Auto) and control output (%) will be displayed on the bottom line. If controlling pH, the reagent will also be displayed. In addition, for the display to be enabled, a measurement must be assigned under Tune Parameters and a relay or analog output must be assigned under Mode.

In Manual, the control output may be adjusted with the up and down arrow keys.

(The ”Info” key function is not available in Manual.)

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

11 Service

(PATH: Menu/Service)

Diagnostics

Service

Calibrate

78

Tech Service

11.1 Enter Service Menu

While in Measurement mode press the b key. Press the m or .

key to navigate to the ”Service”

Menu and press [ENTER]. The available system configuration options are detailed below

11.2 Diagnostics

(PATH: Menu/Service/Diagnostics)

Enter Service Menu as described in section 11.1 ”Enter Service Menu” and press [ENTER] .

This Menu is a valuable tool for troubleshooting and provides diagnostic functionality for the following items: Model/Software Revision, Digital Input, Display, Keypad, Memory, Set Relays,

Read Relays, Set Analog Outputs, Read Analog Outputs.

11.2.1 Model/Software Revision

Essential information for every Service call is the model and software revision number.

This menu shows the part number, model and the serial number of the transmitter. By using the

.

key it is possible to navigate forward through this submenu and get additional information like the current version of firmware implemented on the transmitter (Master V_XXXX and

Comm V_XXXX); and – if an ISM sensor is connected – the version of the sensor firmware

(Sensor FW V_XXX) and sensor hardware (Sensor HW XXXX).

Press [ENTER] to exit from this display.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

11.2.2 Digital Input

The digital Input menu shows the state of the digital inputs. Press [ENTER] to exit from this display.

79

11.2.3 Display

All pixels of the display will be lit for 15 seconds to allow troubleshooting of the display. After

15 seconds the transmitter will return to the normal Measuring mode or press [ENTER] to exit sooner.

11.2.4 Keypad

For keypad diagnostics, the display will indicate which key is pressed. Pressing [ENTER] will return the transmitter to the normal Measuring mode.

11.2.5 Memory

If Memory is selected then the transmitter will perform a RAM and ROM memory test. Test patterns will be written to and read from all RAM memory locations. The ROM checksum will be recalculated and compared to the value stored in the ROM.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

11.2.6 Set Relay

The Set Relays diagnostic menu allows to open or close each relay manually. To access relays

5 and 6, press [ENTER].

0 = open the relay

1 = close the relay

Press [ENTER] to return to Measurement mode.

80

11.2.7 Read Relays

The Read Relays diagnostic menu shows the state of each Relay as defined below. To display

Relays 5 and 6, press [ENTER]. Press [ENTER] again to exit from this display.

0 = Normal

1 = Inverted.

11.2.8 Set Analog Outputs

This menu enables the user to set all analog outputs to any mA value within the 0–22 mA range. Press [ENTER] to exit from this display.

11.2.9 Read Analog Outputs

This menu shows the mA value of the analog Outputs. Press [ENTER] to exit from this display.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 81

11.3 Calibrate

(PATH: Menu/Service/Calibrate)

Enter Service Menu as described in section 11.1 ”Enter Service Menu”, select Calibrate, and press [ENTER].

This menu has the options to calibrate the transmitter and the analog outputs and also allows the unlocking of calibration functionality.

11.3.1 Calibrate Meter (not on ISM version)

The M300 transmitter is factory calibrated within specifications. It is not normally necessary to perform meter re-calibration unless extreme conditions cause an out of spec operation shown by Calibration Verification. Periodic verification/re-calibration may also be necessary to meet Q.A. requirements. Meter calibration can be selected as Resistance (1–5, used for conductivity), Current (used for most dissolved oxygen and dissolved ozone), Voltage,

Rg Diagnostic, Rr Diagnostic (used for pH and 58037221 dissolved oxygen), and Temperature

(used for all measurements).

11.3.1.1 Resistance

The meter is equipped with five (5) internal ranges of measurement on each channel. Each resistance range and temperature is calibrated separately, with each resistance range consisting of a two-point calibration.

Below is a table showing the resistance values for all calibration ranges.

Range:

Resistivity 1

Resistivity 2

Resistivity 3

Resistivity 4

Resistivity 5

Temperature

Point 1

1.0 Mohms

100.0 Kohms

10.0 Kohms

1.0 Kohms

100 Ohms

1000 Ohms

Point 2

10.0 Mohms

1.0 Mohms

100.0 Kohms

10.0 Kohms

1.0 Kohms

3.0 Kohms

Point 3

66 Kohms

It is recommended that both calibration and verification be performed using the M300 Calibrator

Module Accessory (refer to accessory list, in section 15). Instructions on the use of this accessory are provided with the calibrator module.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 82

Navigate to the Calibrate Meter screen and select Channel A or B, and Resistance 1, designating that the transmitter is ready to calibrate the first range resistor. This resistance may be changed, selecting range 1 thru 5. Each resistance range consists of a two-point calibration.

Press [ENTER] to begin the calibration process.

The first text line will ask for the Point 1 resistance value (this will correspond to Resistance 1 value shown on the Calibration Module Accessory). The second text line will show the measured resistance value. When the value stabilizes, press [ENTER] to perform calibration.

The transmitter screen will then prompt the user to enter the value for Point 2, and R1 will display the measured resistance value. When this value stabilizes, press [ENTER] to calibrate this range and bring up a confirmation screen.

Select Yes to save the calibration values and the successful Calibration is confirmed on the display. The transmitter will return to the Measurement mode in approximately 5 seconds.

Once point 1 and 2 are calibrated, return to the Calibrate Meter screen. Move the cursor to change to Resistance 2, designating the second calibration range. Proceed with the two-point calibration process as performed for the first range. This same process must be followed to complete the resistance calibration of all 5 ranges.

11.3.1.2 Temperature

Temperature is performed as a three point calibration. The table above shows the resistance values of these three points.

Navigate to the Calibrate Meter screen and choose Temperature calibration for either Channel A or B.

Press [ENTER] to begin temperature calibration process

The first text line will ask for the Point 1 temperature resistance value (this will correspond to

Temperature 1 value shown on the Calibration Module Accessory). The second text line will show the measured resistance value. When the value stabilizes, press [ENTER] to perform calibration.

The transmitter screen will then prompt the user to enter the value for Point 2, and T2 will display the measured resistance value. When this value stabilizes, press [ENTER] to calibrate this range.

Repeat these steps for Point 3.

Press [ENTER] to bring up a confirmation screen. Select Yes to save the calibration values and the successful Calibration is confirmed on the display.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

The transmitter will return to the Measurement mode in approximately 5 seconds.

83

11.3.1.3 Current

Current Calibration is preformed as a two point calibration.

Navigate to the Calibrate Meter screen and select Channel A or B and Current.

Enter the value for Point 1, in milliamps, of the current source connected to the input. The second display line will show the measured current. Press

[ENTER] to begin the calibration process.

Enter the value for Point 2, in milliamps, of the current source connected to the input. The second display line shows the measured current.

Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes to save the calibration values and the successful Calibration is confirmed on the display. The transmitter will return to the Measurement mode in approximately 5 seconds.

11.3.1.4 Voltage

Voltage Calibration is preformed as a two point calibration.

Navigate to the Calibrate Meter screen and select Channel A or B and Voltage.

Enter the value for Point 1 in, volts, connected to the input. The second display line will show the measured voltage. Press[ENTER] to begin the calibration process.

Enter the value for Point 2, in volts, of the source connected to the input. The second display line shows the measured voltage.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes to save the calibration values and the successful Calibration is confirmed on the display. The transmitter will return to the Measurement mode in approximately 5 seconds.

84

11.3.1.5 Rg Diagnostic

Rg Diagnostic is performed as a two point calibration. Navigate to the Calibrate Meter screen and select Channel A or B and Rg Diagnostic.

Enter the value for Point 1 of the calibration according to the resistor connected across the pH glass electrode measuring input. Press [ENTER] to begin the calibration process.

Enter the value for Point 2 of the calibration according to the resistor connected across the pH glass electrode measuring input.

Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes to save the calibration values and the successful Calibration is confirmed on the display. The transmitter will return to the Measurement mode in approximately 5 seconds.

11.3.1.6 Rr Diagnostics

Rr Diagnostic is performed as a two point calibration. Navigate to the Calibrate Meter screen and select Channel A or B and Rr Diagnostic.

Enter the value for Point 1 of the calibration according to the resistor connected across the pH reference measuring input. Press [ENTER] to begin the calibration process.

Enter the value for Point 2 of the calibration according to the resistor connected across the pH reference measuring input.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes to save the calibration values and the successful Calibration is confirmed on the display. The transmitter will return to the Measurement mode in approximately 5 seconds.

85

11.3.2 Calibrate Analog

Select the Analog Output you wish to calibrate. Each Analog output can be calibrated at 4 and

20 mA.

Connect an accurate milliamp meter to the Analog output terminals and then adjust the five digit number in the display until the milliamp meter reads 4.00 mA and repeat for 20.00 mA.

As the five digit number is increased the output current increases and as the number is decreased the output current decreases. Thus coarse changes in the output current can be made by changing the thousands or hundreds digits and fine changes can be made by changing the tens or ones digits.

Pressing the [ENTER] key after entering both values will bring up a confirmation screen.

Selecting No will discard the entered values, selecting Yes will make the entered values the current ones.

11.3.3 Calibrate Unlock

Select this Menu to configure the CAL Menu (see chapter 7 ”Sensor Calibration”).

Selecting Yes means that Meter calibration menu (see chapter 11.3.1 ”Calibrate Meter”) and Analog Output calibration menu (see chapter 11.3.2 ”Calibrate Analog”) will be selectable under the CAL Menu. Selecting No means that only the Sensor calibration is available under the CAL

Menu. Press [ENTER] after the selection to display a confirmation screen.

11.4 Tech Service

(PATH: Menu/Tech Service)

Note: This Menu is for Mettler Toledo Service personnel use only.

h

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 86

12

(PATH: Info)

Info

Info

Messages

ISM Sensor

Info*

Calibration

Data

Model/Software

Revision

* ISM version only

12.1 Info Menu

Pressing the .

key will display the Info Menu with the options Messages, Calibration Data and

Model/Software Revision.

12.2 Messages

(PATH: Info/Messages)

Enter Info Menu as described in section 12.1 ”Info Menu” and press [ENTER].

The most recent message is displayed. The up and down arrow keys allow scrolling through the last four messages that have occurred.

Clear Messages clears all the messages. Messages are added to the message list when the condition that generates the message first occurs. If all messages are cleared and a message condition still exists and started before the clear then it will not appear in the list. For this message to re-occur in the list the condition must go away and then reappear.

12.3 Calibration Data

(PATH: Info/Calibration Data)

Enter Info Menu as described in section 12.1 ”Info Menu”, select Calibration Data, and press

[ENTER].

The menu displays the calibration constants for each sensor. Use the up and down arrow keys to toggle between channels ”A” and ”B”.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

P = calibration constants for the primary measurement

S = calibration constants for the secondary measurement

Press [ENTER] to exit from this display.

87

12.4 Model/Software Revision

(PATH: Info/Model/Software Revision)

Enter Info Menu as described in section 12.1 ”Info Menu”, select Model/Software Revision, and press [ENTER].

Selecting Model/Software Revision will display the part number, model and the serial number of the transmitter. By using the .

key it is possible to navigate forward through this menu and get additional information like the current version of firmware implemented on the transmitter (Master V_XXXX and Comm V_XXXX); and – if an ISM sensor is connected – the version of the sensor firmware (Sensor FW V_XXX) and sensor hardware (Sensor HW XXXX).

The displayed information is important for any Service call. Press [ENTER] to return to the normal measurement mode.

12.5 ISM Sensor Info (ISM version only)

(PATH: Info/ISM Sensor Info)

Enter Info Menu as described in section 12.1 ”Info Menu”, select ISM Sensor Info, and press

[ENTER].

After plugging in an ISM sensor, the following information about the sensor will be shown in this menu. Use up and down arrows to scroll in the menu.

Type: Type of sensor (e.g. InPro 3250)

Cal Date: Date of the last calibration

Serial-No.: Serial number of the connected Sensor

Part-No.: Part number of the connected Sensor

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

13 Maintenance

13.1 For Technical Support

For technical support and product information for M300 Thornton Transmitters contact:

Mettler-Toledo Thornton, Inc.

36 Middlesex Turnpike

Bedford, MA 01730 USA

Phone: 781-301-8600 or 800-510-PURE

Fax: 781-271-0214

Email: [email protected]

Or: Your local Mettler-Toledo Sales Office or representative.

88

13.2 Front Panel Cleaning

Clean the front panel with a damp soft cloth (water only, no solvents). Gently wipe the surface and dry with a soft cloth.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

14 Troubleshooting

If the equipment is used in a manner not specified by Mettler-Toledo Thornton, Inc., the protection provided by the equipment may be impaired.

Review the table below for possible causes of common problems:

89

Problem

Display is blank.

Incorrect measurement readings.

Measurement readings not stable.

Displayed a is flashing.

Cannot change menu settings.

Possible Cause

– No power to M300.

– Blown fuse.

– LCD display contrast set incorrectly.

– Hardware failure.

– Sensor improperly installed.

– Incorrect units multiplier entered.

– Temperature compensation incorrectly set or disabled.

– Sensor or transmitter needs calibration.

– Sensor or patch cord defective or exceeds recommended maximum length.

– Hardware failure.

– Sensors or cables installed too close to equipment that generates high level of electrical noise.

– Recommended cable length exceeded.

– Averaging set too low.

– Sensor or patch cord defective.

– Setpoint is in alarm condition (setpoint exceeded).

– Alarm, that has been selected (see chapter 8.5.1

“Alarm”), has occurred.

– User locked out for security reasons.

14.1 Changing the Fuse

a

Make sure that the mains cable is unplugged before changing the fuse. This operation should only be carried out by personnel familiar with the transmitter and who are qualified for such work.

If the power consumption of the M300 transmitter is too high or a malfunction leads to a short circuit the fuse will blow. In this case remove the fuse and replace it with one specified in Section 15 “Accessories and Spare Parts”.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

14.2 pH Error Messages / Warning- and Alarm List

14.2.1 pH sensors execpt dual membrane pH electrodes

Warnings

Warning pH slope > 102%

Warning pH Slope < 90%

Description

Slope too big

Slope too small

Warning pH Zero >7.5 pH

Warning pH Zero < 6.5 pH

Zero offset too big

Zero offset too small

Warning pHGls change < 0.3

Glass electrode resistance changed by more than factor 0.3

Warning pHGls change > 3 Glass electrode resistance changed by more than factor 3

Warning pHRef change < 0.3

Warning pHRef change > 3

Reference electrode resistance changed by more than factor

0.3

Reference electrode resistance changed by more than factor 3

Alarms

Watchdog time-out*

Error pH Slope >103%

Description

SW/System fault

Slope too big

Error pH Slope < 80%

Error pH Zero > 8.0 pH

Slope too small

Zero offset too big

Error pH Zero < 6.0 pH Zero offset too small

Error pH Ref Res >150 KΩ** Reference electrode resistance too big (break)

Error pH Ref Res < 2000 Ω** Reference electrode resistance too small (short)

Error pH Gls Res > 2000 MΩ** Glass electrode resistance too big (break)

Error pH Gls Res < 5 MΩ** Glass electrode resistance too small (short)

* ISM sensors only

** According to the parameterization of the transmitter (see chapter 8.5.1 “Alarm”;

PATH: Menu/Configure/Alarm/Clean/Setup Alarm)

90

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

14.2.2 Dual membrane pH electrodes (pH/pNa)

Warnings

Warning pH slope >102%

Warning pH Slope < 90%

Warning pH Zero > 8.0 pH

Warning pH Zero < 6.0 pH

Description

Slope too big

Slope too small

Zero offset too big

Zero offset too small

Warning pHGls change < 0.3* Glass electrode resistance changed by more than factor 0.3

Warning pHGls change > 3* Glass electrode resistance changed by more than factor 3

Warning pNaGls change< 0. 3* Glass electrode resistance changed by more than factor 0.3

Warning pNaGls change > 3* Reference electrode resistance changed by more than factor 3

91

Alarms

Watchdog time-out

Error pH Slope >103%

Description

SW/System fault

Slope too big

Error pH Slope < 80%

Error pH Zero > 9.0 pH

Slope too small

Zero offset too big

Error pH Zero < 5.0 pH Zero offset too small

Error pNa Gls Res > 2000 MΩ* Glass electrode resistance too big (break)

Error pNa Gls Res < 5 MΩ* Glass electrode resistance too small (short)

Error pH Gls Res > 2000 MΩ* Glass electrode resistance too big (break)

Error pH Gls Res < 5 MΩ* Glass electrode resistance too small (short)

* According to the parameterization of the transmitter (see chapter 8.5.1 “Alarm”;

PATH: Menu/Configure/Alarm/Clean/Setup Alarm)

14.2.3 ORP messages

Warnings* Description

Warning ORP ZeroPt > 30 mV Zero offset too big

Warning ORP ZeroPt < – 30 mV Zero offset too small

Alarms*

Watchdog time-out

Error ORP ZeroPt > 60 mV

Description

SW/System fault

Zero offset too big

Error ORP ZeroPt < – 60 mV Zero offset too small

* ISM sensors only

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

14.3 O2 Error Messages / Warning- and Alarm List

Warnings

Warning O

2

Slope < –90 nA

Warning O

2

Slope > –35 nA

Warning O

2

ZeroPt > 0.3 nA

Warning O

2

ZeroPt < –0.3 nA

Description

Slope too big

Slope too small

Zero offset too big

Zero offset too small

Alarms

Watchdog time-out

Error O

2

Slope < –110 nA

Error O

2

Slope > – 30 nA

Error O

2

ZeroPt > 0.6 nA

Error O

2

ZeroPt < – 0.6 nA

Description

SW/System fault

Slope too big

Slope too small

Zero offset too big

Zero offset too small

92

14.4 Cond Error Messages / Warning- and Alarm List

Alarms

Watchdog time-out

Cond Cell open*

Description

SW/System fault

Cell running dry (no measurement solution) or wires are broken

Short circuit caused by sensor or cable Cond Cell shorted*

* According to the parameterization of the transmitter (see chapter 8.5.1 “Alarm”;

PATH: Menu/Configure/Alarm/Clean/Setup Alarm)

14.5 O

2

(I) Error Messages / Warning- and Alarm List

(Thornton Models only)

Warnings

Warning DO Slope < – 460 nA

Warning DO Slope > – 250 nA

Warning DO ZeroPt > 0.5 nA

Warning DO ZeroPt < – 0.5 nA

Description

Slope too big

Slope too small

Zero offset too big

Zero offset too small

Alarms

Watchdog time-out

Error Install O2 Jumper

Error DO Slope < – 525 nA

Error DO Slope > – 220 nA

Error DO ZeroPt > 1.0 nA

Error DO ZeroPt < – 1.0 nA

Description

SW/System fault

Wrong installation of jumper

Slope too big

Slope too small

Zero offset too big

Zero offset too small

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

14.6 O

2

(V) Error Messages / Warning- and Alarm List

(Thornton Models only)

Warnings

Warning DO Slope >1.50

Warning DO Slope < 0.65

Warning DO ZeroPt >15 μV

Warning DO ZeroPt < –15 μV

Description

Slope too big

Slope too small

Zero offset too big

Zero offset too small

Alarms

Watchdog time-out

Warning DO Slope < 2.00

Warning DO Slope < 0.25

Warning DO ZeroPt > 30 μV

Warning DO ZeroPt < –30 μV

Description

SW/System fault

Slope too big

Slope too small

Zero offset too big

Zero offset too small

14.7 Ozone Error Messages / Warning- and Alarm List

(Thornton Models only)

Warnings

Warning O

3

Slope >1.83 nA

Warning O

3

Slope < 0.73 nA

Warning O

3

ZeroPt > 0.5 nA

Warning O

3

ZeroPt < – 0.5 nA

Description

Slope too big

Slope too small

Zero offset too big

Zero offset too small

Alarms

Watchdog time-out

Error O

3

Slope > 2.75 nA

Error O

3

Slope < 0.65 nA

Error O

3

ZeroPt >1.0 nA

Error O

2

ZeroPt < –1.0 nA

Description

SW/System fault

Slope too big

Slope too small

Zero offset too big

Zero offset too small

93

14.8 Warning- and Alarm indication on the display

14.8.1 Warning indication

If there are conditions, that generate a warning, the message will be recorded through the menu

Messages (see chapter 12.1 “Messages”; PATH: Info/Messages). According to the parameterisation of the transmitter the hint ”Failure – Press Enter” will be shown at line 4 of the display if a warning or alarm has occurred (see chapter 8.6. “Display”;

PATH: Menu/Configure/Display/Measurement).

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 94

14.8.2 Alarm indication

Alarms will be shown in the display by a flashing symbol a and recorded through the menu

Messages (see chapter 12.1 “Messages”; PATH: Info/Messages).

Furthermore the detection of some alarms can be activated or deactivated (see chapter 8.5

“Alarm/Clean”; PATH: Menu/Configure/Alarm/Clean) for an indication on the display. If one of these alarms occurs and the detection has been activated, the flashing symbol a will be shown on the display. The message will be recorded through the menu Messages

(see chapter 12.1 “Messages”; PATH: Info/Messages).

Alarms, that are caused by a violation of the limitation of a setpoint or the range (see chapter

8.4 “Setpoints”; PATH: Menu/Configure/Setpoint) will be shown by a flashing symbol a and recorded through the menu Messages (see chapter 12.1 “Messages”; PATH: Info/Messages).

According to the parameterisation of the transmitter the hint ”Failure – Press Enter” will be shown at line 4 of the display if a warning or alarm has occurred (see chapter 8.6. “Display”;

PATH: Menu/Configure/Display/Measurement).

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

15 Accessories and Spare Parts

Please contact your local Mettler-Toledo Sales office or representative for details for additional accessories and spare parts.

For M300 Thornton

Description

Pipe Mount Kit for 1/2DIN models

Panel Mount Kit for 1/2DIN models

Adaptor Panel – M300 to 200/2000 cutout

M300 Conductivity Calibrator Module

Replacement power fuse 5x20 mm, 1 A, 250 V, time lag, Littlefuse or Hollyland

Terminal blocks for M300

Order no.

52 500 212

52 500 213

58 083 300

58 082 300

58 091 326

52 121 504

95

For M300

Description

Pipe Mount Kit for 1/2DIN models

Panel Mount Kit for 1/2DIN models

Protective Hood for 1/2DIN models

Terminal blocks for M300, M400

Order no.

52 500 212

52 500 213

52 500 214

52 121 504

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

16 Specifications

16.1 General specifications

Conductivity/resistivity Specifications

Range 0.01 cm –1 constant sensor 0.002 to 200 µS/cm (5000 Ω x cm to 500 MΩ x cm)

Range 0.1 cm –1 constant sensor 0.02 to 2000 µS/cm (500 Ω x cm to 50 MΩ x cm)

Range 10 cm –1 constant sensor

Display range for 2-e sensor

Display range for 4-e sensor

10 to 40,000 µS/cm (25 Ω x cm to 100 KΩ x cm)

0 to 40,000 mS/cm (25 Ω x cm to 100 MΩ x cm)

0.01 to 650 mS/cm (1.54 Ω x cm to 0.1MΩ x cm)

Chemical concentration curves

NaCl: 0–26% @ 0°C to 0–28% @ +100 °C

NaOH: 0–12% @ 0°C to 0–16% @ +40 °C to 0–6% @ +100 °C

HCl: 0–18% @ –20 °C to 0–18% @ 0 °C to 0-5% @ +50 °C

HNO3: 0–30% @ –20 °C to 0–30% @ 0 °C to 0-8% @ +50 °C

H2SO4: 0–26% @ –12 °C to 0–26% @ +5 °C to 0–9% @ +100 °C

H3PO4: 0–35% @ +5 °C to +80 °C

TDS ranges

Temperature input*

Temperature measuring range

NaCl, CaCO3

Pt1000

– 40 to + 200.0 °C (– 40 to 392 °F)

Sensor maximum distance

Analogue 2-e: 61 m (200 ft),

Analogue 4-e: 15 m (50 ft),

ISM 2-e: 90 m (300 ft)

ISM 4-e: 80 m (260 ft),

Cond/Res resolution

Cond/Res accuracy**

Cond/Res repeatability**

Temperature resolution

Temperature accuracy**

Temperature repeatability** pH Specifications pH range mV range

Temperature input*

Temperature measuring range

Sensor maximum distance pH resolution pH accuracy** mV resolution mV accuracy

Temperature resolution

Temperature accuracy**

Auto / 0.001 / 0.01 / 0.1 / 1 (can be selected)

± 0.5% of reading or 0.25Ω, whichever is greater,

± 0.25% of reading or 0.25 ohm, whichever is greater

Auto / 0.001 / 0.01 / 0.1 / 1 °C (°F) (can be selected)

± 0.25°C (± 0.45 °F)

± 0.13°C (± 0.23 °F)

–2.00 to 16.00 pH

–1500 to 1500 mV

Pt1000 (Pt100 with adapter)

– 30 to 130 °C (– 22 to 266 °F)

Analogue: 10 to 20 m (33 to 65 ft) depending on sensor

ISM: 80 m (260 ft) auto / 0.01 / 0.1 / 1 (can be selected)

± 0.02 pH

1 mV

± 1 mV

Auto / 0.001 / 0.01 / 0.1 / 1 °C (°F), (can be selected)

± 0.25°C (± 0.45 °F)

* Not requried on ISM sensors

** For analogue input signal (ISM signal causes no additional error)

96

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

Available Buffer Sets:

MT-9 buffers, MT-10 buffers, NIST Technical Buffers,

NIST Standard Buffers (DIN 19266:2000–01), JIS Z 8802 buffers, Hach buffers, CIBA (94) buffers, Merck Titrisols-Reidel Fixanals, WTW buffers

Dual membrane electrodes pH buffers (pH/pNa)

Mettler-pH/pNa buffers (Na+ 3.9M)

Dissolved Oxygen Specifications

Measuring current range

Concentration range

Temperature input*

Temperature measuring range

Sensor maximum distance

DO resolution

DO accuracy**

Temperature resolution

Temperature accuracy**

Polarization voltage

Dissolved Ozone Specifications

Ozone range

Ozone resolution

Relative accuracy

Temperature input

0 to 900 nA

0.00 to 50.00 ppm (mg/l)

NTC 22 kΩ

–10 to 80 °C (14 to 176 °F)

Analogue: 20 m ( 65 ft)

ISM: 80 m (260 ft)

Auto / 0.001 / 0.01 / 0.1 / 1, (can be selected)

± 0.5% of full scale reading

Auto / 0.001 / 0.01 / 0.1 / 1 °C (°F), (can be selected)

± 0.25 °C (± 0.45 °F)

-674 mV (for analog sensors)

0 – 5,000 ppb, 0 – 5 ppm

1 ppb, 0.001 ppm

± 2% of reading or ± 3 ppb, system

Pt1000

* Not requried on ISM sensors

** For analogue input signal (ISM signal causes no additional error)

97

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 h

16.2 Electrical specifications for

1/2DIN and 1/4DIN versions

Power requirements

Frequency

Analog output signals

Measurement Error through analog outputs

Analog output configuration

Load

Connection terminals

Digital communication

PID process controller

Cycle time

Connection terminals

Digital Input

Mains power fuse

Relays

Alarm Relay delay

Keypad

Display

100 to 240 VAC or 20 to 30 VDC, 10 VA;

AWG 14 < 2.5 mm 2

50 to 60 Hz

4 (2 for single channel version) 0/4 to 22 mA outputs, galvanically isolated from input and from earth/ground

< ± 0.05 mA over 1 to 22 mA range,

< ± 0.1 mA over 0 to 1 mA range

Linear, Bi-Linear, Logarithmic, Autoranging

Max. 500 Ω

Detachable screw terminals

USB port, Type B connector

Pulse length, pulse frequency or analog control

Ca 1 sec.

Detachable screw terminals

1 (2 for dual channel version) with swtiching limits

0.00 VDC to 1.00 VDC for low level

2.30 VDC to 30.00 VDC for high level

1.0 A slow blow type FC

– 2-SPDT mechanical 250 VAC, 30 VDC, 3 Amps

– 2-SPST mechanical rated at 250 VAC, 3 Amps

(dual channel only)

– 2-Reed 250 VAC or DC, 0.5 A, 10 W

0–999 s

5 tactile feedback keys

Backlit LCD, four-line

NOTE: This is a 4-wire-product with an active 4–20 mA analog output.

Please do not supply to Pin1–Pin6 of TB2.

98

16.3 Mechanical specifications for 1/4DIN version

Dimensions (housing – H x W x D)* 96 x 96 x 140 mm (1/4DIN model)

Front bezel – (H x W) 102 x 102 mm

Max. depth 125 mm (excludes plug-in connectors)

Weight

Material

Ingress rating

0.6 kg (1.5 lb)

ABS/polycarbonate

IP 65 (front)/IP 20 (housing)

* H = Height, W = Width, D = Depth

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

16.4 Mechanical specifications for 1/2DIN version

Dimensions (housing – H x W x D)* 144 x 144 x 116 mm

Front bezel – H x W

Max. D – panel mounted

Weight

Material

Ingress rating

150 x 150 mm

87 mm (excludes plug-in connectors)

0.95 kg (2 lb)

ABS/polycarbonate

IP 65 (when back cover is attached)

* H = Height, W = Width, D = Depth

16.5 Environmental specifications for

1/2DIN and 1/4DIN versions

Storage temperature – 40 to 70 °C (– 40 to 158 °F)

Ambient temperature operating range –10 to 50 °C (14 to 122 °F)

Relative humidity 0 to 95% non-condensing

Emissions

UL Electrical Environment

According to EN55011 Class A

Installation (overvoltage) category II

99

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

17 Default tables

17.1

Parameter

Alarm

Clean

Language

Passwords

Lockout

Display

Analog Out

All analog out

M300 ISM (Single-Channel Instruments)

Unit

All Relays (Unless

Otherwise Specified)

Administrator

Operator

Delay

Hysteresis

State

Hold Mode*

Yes/No

Line 1

Line 2

1

2

Line 3

Line 4

Mode

Type

Alarm

Hold Mode

Sub parameter

Relay

Power Failure

Software Failure

Rg Diagnostics

Rr Diagnostics

Cond Cell Open

Cond Cell Shorted

Disconnect ChA

Hold Mode*

Delay

Hysteresis

State

Relay

Hold Mode*

Interval

Clean time

State

Delay

Hysteresis

1

0

No

Last

Inverted

1

0

0

Last

0

0

Normal

No

No

No

No

Value

2

No

No

English

00000

00000

10 a b

5

Normal

Last

No c (not available) d (not available) a b

4–20 mA

Normal

Off

Last Value

Sec

Hrs

Sec

Sec

%

100

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

Parameter

Conductivity

Resitivity

Sub parameter

Value 4 mA

Value

0.1

10

O pH

Temperature

Set Point 1

Conductivity

Resitivity

O

2

2 pH

Relay 3

Set Point 2

Value 20 mA

Value 4 mA

Value 20 mA

Value 4 mA

Value 20 mA

Value 4 mA

Value 20 mA

Measurement

Type

High Value

Low Value

High Value

Low Value

High Value

Low Value

Set point

Measurement

Type

High Value

Low Value

Set Point Relay 4

Resolution

Conductivity

Resistivity

O

2

Compensation Standard pH

V polarisation**

CalPres

ProcPres

ProcCalPres

Salinity

Humidity

Drift Control

IP

STC

FixCalTemp pH Buffer

Cal info slope

Cal info offset

* For analogue output signal if relay is switched

** Not adjustable

–675

759.8

759.8

CalPres

0.0

100

Auto

7.0

0.000

No

Mettler-9

[%]

[pH]

10

20

0

0

0

0

0

100

2

12

0

100 a

Off

0

0

0

1 b

Off

50

0

12

2

Auto

Italics = default values if resitivity instead of coductivity is chosen.

µS/cm

MΩ-cm

µS/cm

MΩ-cm

%sat

%sat pH pH

Unit

µS/cm

MΩ-cm

µS/cm

MΩ-cm

%sat

%sat pH pH

°C

°C

°C

°C mV mmHg mmHg g/kg

% pH pH/°C

101

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

17.2

Parameter

Alarm

Clean

Language

Passwords

M300 ISM (Dual-Channel Instruments)

Sub parameter

Relay

Power Failure

Software Failure

Rg Diagnostics

Rr Diagnostics

Cond Cell Open

Cond Cell Shorted

Disconnect ChA

Disconnect CHB

Hold Mode*

Delay

Hysteresis

State

Relay

Hold Mode*

Interval

Clean time

State

Delay

Hysteresis

Administrator

Operator

Last

1

0

Inverted

1

Last

0

0

Normal

0

0

English

00000

00000

No

No

No

No

No

No

Value

2

No

No

All Relays (Unless

Otherwise Specified)

Delay 10

Lockout

Display

Analog Out

All analog out

Conductivity

Resitivity

O

2

Hysteresis

State

Hold Mode*

Yes/No

Line 1

Line 2

Line 3

Line 4

3

4

1

2

Mode

Type

Alarm

Hold Mode

Value 4 mA

Value 20 mA

Value 4 mA

Value 20 mA a b c d

5

Normal

Last

No c d a b

4–20 mA

Normal

Off

Last Value

0.1

10

10

20

0

100

Unit

Sec

Hrs

Sec

Sec

%

µS/cm

MΩ-cm

µS/cm

MΩ-cm

%sat

%sat

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

102

Transmitter Multiparameter M300

52 121 389

Transmitter M300 103

Parameter pH

Temperature

Set Point 1

Conductivity

Resitivity

O

2 pH

Relay 3

Set Point 2

Conductivity

Resitivity

O

2 pH

Relay 4

Resolution

Set Point 3

Set Point 4

Sub parameter

Value 4 mA

Value 20 mA

Value 4 mA

Value 20 mA

Measurement

Type

High Value

Low Value

High Value

Low Value

High Value

Low Value

Set point

Measurement

Type

High Value

Low Value

High Value

Low Value

High Value

Low Value

Set Point

Measurement

Type

Relay

Measurement

Type

Relay

Conductivity

Resistivity

O

2 pH

Compensation

V polarisation**

CalPres

ProcPres

ProcCalPres

Salinity

Humidity

Drift Control

IP

STC

FixCalTemp pH Buffer

Cal info slope

Cal info offset

Standard

–675

759.8

759.8

CalPres

0.0

100

Auto

7.0

0.000

No

Mettler-9

[%]

[pH]

* For analogue output signal if relay is switched ** Not adjustable

Value

2

12

0

100 a

Off

0

0

0

0

50

0

12

0

1

0

0

0

0 c

Off

50

0

12

0

2

Auto

_(none)

Off

_(none)

_(none)

Off

_(none)

Italics = default values if resitivity instead of coductivity is chosen.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Unit pH pH

°C

°C

µS/cm

MΩ-cm

µS/cm

MΩ-cm

%sat

%sat pH pH

µS/cm

MΩ-cm

µS/cm

MΩ-cm

%sat

%sat pH pH mV mmHg mmHg g/kg

% pH pH/°C

Transmitter Multiparameter M300

52 121 389

Transmitter M300

17.3

Parameter

Alarm

Clean

Language

Passwords

Lockout

Display

Cal constants

Analog Out

All analog out

Conductivity

Resitivity

Temperature

M300 Conductivity (Single-Channel Instruments)

Unit

All Relays (Unless

Otherwise Specified)

Administrator

Operator

Delay

Hysteresis

State

Hold Mode*

Yes/No

Line 1

Line 2

Line 3

Line 4

Cond/Res

Sub parameter

Relay

Power Failure

Software Failure

Cond Cell Open

Cond Cell Shorted

Hold Mode*

Delay

Hysteresis

State

Relay

Hold Mode*

Interval

Clean time

State

Delay

Hysteresis

Temperature

1

2

Mode

Type

Alarm

Hold Mode

Value 4 mA

Value 20 mA

Value 4 mA

Value 20 mA

Value

2

No

No

No

No

Last

1

0

Inverted

1

Last

0

0

0

0

Normal

English

00000

00000

10

5

Normal

Last

No a (Conductivity) b (Temperature) c (not available) d (not available)

M = 0.1

A = 0.0

M = 1.0,

A = 0.0

a (Resitivity) b (Temperature)

4–20 mA

Normal

Off

Last Value

0.1

10

10

20

0

100

Sec

Hrs

Sec

Sec

%

S/cm

°C cm -1

Ω

Ω

µS/cm

MΩ-cm

µS/cm

MΩ-cm

°C

°C

104

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

Parameter

Set Point 1

Sub parameter

Measurement

Type

High Value

Low Value

Relay 3

Set Point 2

Set point

Measurement

Type

High Value

Low Value

Set Point Relay 4

Resolution

Conductivity

Resistivity

Compensation

* For analogue output signal if relay is switched

1 b

0

0

Off

0

0

2

Auto

0

0

Value a

Off

Standard

Italics = default values if resitivity instead of coductivity is chosen.

Unit

µS/cm

MΩ-cm

µS/cm

MΩ-cm

°C

°C

17.4

Parameter

Alarm

Clean

Language

Passwords

Lockout

M300 O

2

(Single-Channel Instruments)

Unit

All Relays (Unless

Otherwise Specified)

Sub parameter

Relay

Power Failure

Software Failure

Hold Mode*

Delay

Hysteresis

State

Relay

Hold Mode*

Interval

Clean time

State

Delay

Hysteresis

Administrator

Operator

Delay

Hysteresis

State

Hold Mode*

Yes/No

0

0

Normal

0

0

English

00000

00000

Value

2

No

No

Last

1

0

Inverted

1

Last

10

5

Normal

Last

No

Sec

Hrs

Sec

Sec

%

105

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

Parameter

Display

Cal constants

Analog Out

All analog out

O2

Temperature

Set Point 1

Relay 3

Set Point 2

Sub parameter

Line 1

Line 2

Line 3

Line 4

O2

Temperature

1

2

Mode

Type

Alarm

Hold Mode

Value 4 mA

Value 20 mA

Value 4 mA

Value 20 mA

Measurement

Type

High Value

Low Value

Set point

Measurement

Type

High Value

Low Value

Set Point Relay 4

Resolution

O2 V polarisation**

CalPres

ProcPres

ProcCalPres

Salinity

Humidity

* For analogue output signal if relay is switched

** Not adjustable

Value a (O2) b (Temperature) c (not available) d (not available)

S = -70.00

A = 0.0

M = 1.0

A = 0.0

0

2

Off

0

Auto

–675

1 b a

Off

50

0 a (O2) b (Temperature)

4–20 mA

Normal

Off

Last Value

0

100

0

100

759.8

759.8

CalPres

0.0

100 nA nA

Ω

Unit

% sat

°C

% sat

% sat

°C

°C

% sat

% sat

°C

°C mV mmHg mmHg g/kg

%

106

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

17.5 M300 pH (Single-Channel Instruments)

Parameter

Alarm

Clean

Language

Passwords

All Relays (Unless

Otherwise Specified)

Lockout

Display

Cal constants

Administrator

Operator

Delay

Hysteresis

State

Hold Mode*

Yes/No

Line 1

Line 2

Line 3

Line 4 pH

Sub parameter

Relay

Power Failure

Software Failure

Rg diagnostics

Rr diagnostics

Hold Mode*

Delay

Hysteresis

State

Relay

Hold Mode*

Interval

Clean time

State

Delay

Hysteresis

Analog Out

All analog out pH

Temperature

Temperature

1

2

Mode

Type

Alarm

Hold Mode

Value 4 mA

Value 20 mA

Value 4 mA

Value 20 mA

Unit

Sec

Hrs

Sec

Sec

% pH

°C pH pH

°C

°C

10

5

Normal

Last

No a (pH) b (Temperature) c (not available) d (not available)

S = 100

Z = 7.0

M = 1.0

A = 0.0

a (pH) b (Temperature)

4–20 mA

Normal

Off

Last Value

2

12

0

100

0

0

1

Last

Normal

0

0

English

00000

00000

Value

2

No

No

No

No

Last

1

0

Inverted

% pH

Ω

107

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

Parameter

Set Point 1

Relay 3

Set Point 2

Sub parameter

Measurement

Type

High Value

Low Value

Set point

Measurement

Type

High Value

Low Value

Set Point Relay 4

Resolution pH Drift Control

IP

STC

Fix CalTemp pH Buffer

Cal info slope

Cal info offset

* For analogue output signal if relay is switched

0

1 b

Off

0

0

2

Value a

Off

12

Auto

Auto

7.0

0.000

No

Mettler-9

[%]

[pH]

Unit pH pH

°C

°C pH/°C

108

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

17.6

Parameter

Alarm

Clean

Language

Passwords

Lockout

Display

Cal Constants

M300 Multiparameter (Dual-Channel Instruments)

Unit

All Relays (Unless

Otherwise Specified)

Administrator

Operator

Delay

Hysteresis

State

Hold Mode*

Yes/No

Line 1

Line 2

Line 3

Line 4

Cond/Res

Sub parameter

Relay

Power Failure

Software Failure

Rg Diagnostics

Rr Diagnostics

Cond Cell Open

Cond Cell Shorted

Hold Mode*

Delay

Hysteresis

State

Relay

Hold Mode*

Interval

Clean time

State

Delay

Hysteresis

O2

O2(I)***

O2(V)*** pH

O3***

Temperature

10 a b c d

5

Normal

Last

No

M = 0.1

A = 0.0

S = –70.00

Z = 0.00

S = -350.00

Z = 0.00

S = 1.000

Z = 0.000

S = 100

Z = 7.0

S = –1.000

M = 0.000

M = 1.0

A = 0.0

No

No

No

Value

2

No

No

1

0

No

Last

Inverted

1

0

0

Last

0

0

Normal

English

00000

00000

Sec

Hrs

Sec

Sec

% nA nA

µV cm -1

Ω nA nA

% pH nA

Ω

109

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

Parameter

Analog Out

All analog out

2

3

Sub parameter

1

4

Mode

Type

Alarm

Hold Mode

Value 4 mA

Conductivity

Resitivity

Value 20 mA

O

2 pH

Value 4 mA

Value 20 mA

Value 4 mA

Value 20 mA

O2(I)***

O2(V)***

Value 4 mA

Value 20 mA

Value 4 mA

Value 20 mA

Dissolved Ozone*** Value 4 mA

Temperature

Set Point 1

Value 20 mA

Value 4 mA

Value 20 mA

Measurement

Type

Conductivity

Resitivity

High Value

Low Value

O

2 pH

High Value

Low Value

High Value

Low Value

High Value O2(I)***

O2(V)***

Low Value

High Value

Low Value

Dissolved Ozone*** High Value

Relay 3

Set Point 2

Low Value

Set point

Measurement

Type

Conductivity

Resitivity

High Value

Low Value

0

0

0

0

0

0

10

20

100

0.000

20.00

0

100 a

Off

0

100

2

12

0

100

0

0

0

50

0

12

0

40.00

0.000

0.000

0.000

0.000

0.000

1 c

Off

Value a b c d

4–20 mA

Normal

Off

Last Value

0.1

10

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Unit

110

Transmitter Multiparameter M300

52 121 389

µS/cm

MΩ-cm

µS/cm

MΩ-cm ppb ppb ppm

°C

°C

%sat

%sat pH pH ppb ppb ppb

µS/cm

MΩ-cm

µS/cm

MΩ-cm

%sat

%sat pH pH ppb ppb ppb ppb ppb ppb

µS/cm

MΩ-cm

µS/cm

MΩ-cm

Transmitter M300

Parameter

O

2 pH

Sub parameter

High Value

Low Value

High Value

O2(I)***

Set Point 4

Low Value

High Value

Low Value

High Value O2(V)***

Relay 4

Resolution

Low Value

Dissolved Ozone*** High Value

Low Value

Set Point

Set Point 3 Measurement

Type

Relay

Measurement

Type

Relay

Conductivity

Resistivity

O

2

Compensation pH

V polarisation**

CalPres

ProcPres

ProcCalPres

Salinity

Humidity

Drift Control

IP

STC

FixCalTemp pH Buffer

Cal info slope

Cal info offset

Standard

* For analogue output signal if relay is switched

** Not adjustable

*** Thornton models only

–675

759.8

759.8

CalPres

0.0

100

Auto

7.0

0.000

No

Mettler-9

[%]

[pH]

2

Auto

_(none)

Off

_(none)

_(none)

Off

_(none)

Value

50

0

12

0

40.00

0.000

0.000

0.000

0.000

0.000

Italics = default values if resitivity instead of coductivity is chosen.

pH ppb ppb ppb

Unit

%sat

%sat pH ppb ppb ppb mV mmHg mmHg g/kg

% pH pH/°C

111

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

17.7

Parameter

Alarm

Clean

Language

Passwords

Lockout

Display

Cal Constants

Analog Out

All analog out

Conductivity

Resitivity

M300 Conductivity (Dual-Channel Instruments,

Thornton Models only)

Unit

All Relays (Unless

Otherwise Specified)

Administrator

Operator

Delay

Hysteresis

State

Hold Mode*

Yes/No

Line 1

Line 2

Line 3

Line 4

Cond/Res

Sub parameter

Relay

Power Failure

Software Failure

Cond Cell Open

Cond Cell Shorted

Hold Mode*

Delay

Hysteresis

State

Relay

Hold Mode*

Interval

Clean time

State

Delay

Hysteresis

Temperature

3

4

1

2

Mode

Type

Alarm

Hold Mode

Value 4 mA

Value 20 mA

Value

2

No

No

No

No

Last

1

0

Inverted

1

Last

0

0

0

0

Normal

English

00000

00000

10

5

Normal

Last

No a (Resistivity) b (Temperature) c (Resistivity) d (Temperature)

M = 0.1

A = 0.0

M = 1.0

A = 0.0

a (Resistivity) b (Temperature) c (Resistivity) d (Temperature)

4–20 mA

Normal

Off

Last Value

0.1

10

10

20

Sec

Hrs

Sec

Sec

%

Ω-cm

°C

Ω-cm

°C cm -1

Ω

Ω

µS/cm

MΩ-cm

µS/cm

MΩ-cm

112

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

Parameter

Temperature

Set Point 1

Conductivity

Resitivity

Relay 3

Set Point 2

Conductivity

Resitivity

Sub parameter

Value 4 mA

Value 20 mA

Measurement

Type

High Value

Low Value

Set point

Measurement

Type

High Value

Low Value

Relay 4

Resolution

Set Point 3

Set Point 4

Set Point

Measurement

Type

Relay

Measurement

Type

Relay

Conductivity

Resistivity

Compensation

* For analogue output signal if relay is switched

0

0

0

0

2

Auto

_(none)

Off

_(none)

_(none)

Off

_(none)

0

0

0

0

1 c

Off

Value

0

100 a (Resistivity)

Off

Standard

Italics = default values if resitivity instead of coductivity is chosen.

Unit

°C

°C

µS/cm

MΩ-cm

µS/cm

MΩ-cm

µS/cm

MΩ-cm

µS/cm

MΩ-cm

113

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 114

18 Warranty

METTLER TOLEDO warrants this product to be free from significant deviations in material and workmanship for a period of one year from the date of purchase. If repair is necessary and not the result of abuse or misuse within the warranty period, please return by freight pre-paid and amendment will be made without any charge. METTLER TOLEDO”s Customer Service Dept. will determine if the product problem is due to deviations or customer abuse. Out-of-warranty products will be repaired on an exchange basis at cost.

The above warranty is the only warranty made by METTLER TOLEDO and is lieu of all other warranties, expressed or implied, including, without limitation, implied warranties of merchantability and fitness for a particular purpose. METTLER TOLEDO shall not be liable for any loss, claim, expense or damage caused by, contributed to or arising out of the acts or omissions of the Buyer or Third Parties, whether negligent or otherwise. In no event shall METTLER TOLEDO’s liability for any cause of action whatsoever exceed the cost of the item giving rise to the claim, whether based in contract, warranty, indemnity, or tort (including negligence).

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 115

19 Certificate

Mettler-Toledo Thornton, Inc., 36 Middlesex Turnpike, Bedford, MA 01730, USA has obtained

Underwriters Laboratories’ listing for M300 Model Transmitters. They bear the cULus Listed mark, signifying that the products have been evaluated to the applicable ANSI/UL and CSA Standards for use in the U.S. and Canada.

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

20 Buffer tables

M300 transmitters have the ability to do automatic pH buffer recognition. The following tables show different standard buffers that are automatically recognized.

116

20.1 Standard pH buffers

60

65

70

40

45

50

55

75

80

85

90

95

Temp (°C)

20

25

30

35

0

5

10

15

20.1.1 Mettler-9

1.98

1.98

1.98

1.98

1.98

1.98

1.99

pH of buffer solutions

2.03

4.01

2.02

2.01

4.01

4.00

2.00

2.00

2.00

1.99

1.99

4.00

4.00

4.01

4.01

4.02

1.99

2.00

2.00

2.00

2.00

4.19

4.22

4.26

4.30

4.35

4.03

4.04

4.06

4.08

4.10

4.13

4.16

6.97

6.97

6.97

6.98

6.98

6.99

7.00

7.12

7.09

7.06

7.04

7.02

7.00

6.99

6.98

7.02

7.04

7.06

7.09

7.12

9.06

9.03

8.99

8.96

8.93

8.90

8.88

9.52

9.45

9.38

9.32

9.26

9.21

9.16

9.11

8.85

8.83

8.81

8.79

8.77

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

85

90

95

65

70

75

80

50

55

60

30

35

40

45

15

20

25

Temp (°C)

0

5

10

20.1.2 Mettler-10

1.99

1.98

1.99

2.00

2.00

2.00

2.00

1.99

1.99

1.98

1.98

1.98

1.98

1.98

pH of buffer solutions

2.03

4.01

2.02

2.01

4.01

4.00

2.00

2.00

2.00

4.00

4.00

4.01

4.01

4.02

4.03

4.04

4.06

4.08

4.10

4.13

4.16

4.19

4.22

4.26

4.30

4.35

20.1.3 NIST Technical Buffers

80

85

90

60

65

70

75

95

40

45

50

55

20

25

30

35

Temp (°C)

0

5

10

15

1.72

1.73

1.74

1.75

1.765

1.78

1.79

1.805

pH of buffer solutions

1.67

4.00

1.67

1.67

1.67

4.00

4.00

4.00

1.675

1.68

1.68

1.69

1.69

1.70

1.705

1.715

4.00

4.005

4.015

4.025

4.03

4.045

4.06

4.075

4.085

4.10

4.13

4.14

4.16

4.18

4.21

4.23

6.97

6.98

6.99

7.01

7.03

7.05

7.08

7.11

7.115

7.085

7.06

7.04

7.015

7.00

6.985

6.98

6.975

6.975

6.97

6.97

6.99

6.98

6.97

6.97

6.97

6.98

6.98

7.12

7.09

7.06

7.04

7.02

7.00

6.99

7.00

7.02

7.04

7.06

7.09

7.12

10.32

10.25

10.18

10.12

10.07

10.01

9.97

9.93

9.89

9.86

9.83

10.65

10.52

10.39

10.26

10.13

10.00

9.87

9.74

9.61

9.48

9.35

13.42

13.21

13.01

12.80

12.64

12.46

12.30

12.13

11.99

11.84

11.71

11.57

11.45

117

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300 h

118

20.1.4 NIST standard buffers (DIN and JIS 19266: 2000–01)

50

55

60

30

35

40

45

70

80

90

95

15

20

25

Temp (°C)

0

5

10 pH of buffer solutions

1.668

1.670

1.672

1.676

1.680

1.685

1.694

1.697

1.704

1.712

1.715

1.723

1.743

1.766

1.792

1.806

4.004

4.001

4.001

4.003

4.008

4.015

4.028

4.036

4.049

4.064

4.075

4.091

4.126

4.164

4.205

4.227

6.950

6.922

6.900

6.880

6.865

6.853

6.841

6.837

6.834

6.833

6.834

6.836

6.845

6.859

6.877

6.886

9.392

9.331

9.277

9.228

9.184

9.144

9.095

9.076

9.046

9.018

8.985

8.962

8.921

8.885

8.850

8.833

NOTE: The pH(S) values of the individual charges of the secondary reference materials are documented in a certificate of an accredited laboratory. This certificate is supplied with the respective buffer materials. Only these pH(S) values shall be used as standard values for the secondary reference buffer materials. Correspondingly, this standard does not include a table with standard pH values for practical use. The table above only provides examples of pH(PS) values for orientation.

20.1.5 Hach buffers

Buffer values up to 60 °C as specified by Bergmann & Beving Process AB.

35

40

45

50

55

60

Temp (°C)

20

25

30

0

5

10

15

4.02

4.03

4.05

4.06

4.07

4.09

pH of buffer solutions

4.00

4.00

4.00

4.00

4.00

4.01

4.01

6.98

6.98

6.98

6.98

6.98

6.99

7.14

7.10

7.04

7.04

7.02

7.00

6.99

9.92

9.88

9.85

9.82

9.79

9.76

10.30

10.23

10.11

10.11

10.05

10.00

9.96

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

85

90

95

65

70

75

80

50

55

60

30

35

40

45

15

20

25

Temp (°C)

0

5

10

* Extrapolated

20.1.6 Ciba (94) buffers

2.07*

2.07

2.04*

2.02

2.03*

2.04

2.05*

2.06

2.06

2.07

2.06

2.06

2.05

2.08

pH of buffer solutions

2.04

4.00

2.09

2.07

4.02

4.00

2.08

2.09

2.08

4.00

4.01

4.02

4.00

4.01

4.02

4.03

4.04

4.05

4.10

4.10*

4.11

4.13*

4.15

4.17*

4.20

4.22*

6.96

6.95

6.94

6.93

6.93

6.91

6.93

7.10

7.08

7.05

7.02

6.98

6.98

6.92*

6.92

6.92*

6.93

6.95*

6.97

6.99*

60

65

70

40

45

50

55

75

80

85

90

95

Temp (°C)

20

25

30

35

0

5

10

15

20.1.7 Merck Titrisole, Riedel-de-Haën Fixanale

2.00

2.00

2.00

2.00

2.00

2.00

2.01

pH of buffer solutions

2.01

4.05

2.01

2.01

4.05

4.02

2.00

2.00

2.00

2.00

2.00

4.01

4.00

4.01

4.01

4.01

2.01

2.01

2.01

2.01

2.01

4.00

4.00

4.00

4.00

4.00

4.01

4.01

4.00

4.00

4.00

4.00

4.00

6.95

6.95

6.95

6.95

6.96

6.96

6.96

7.13

7.07

7.05

7.02

7.00

6.98

6.98

6.96

6.96

6.97

6.98

7.00

7.02

8.85

8.82

8.79

8.76

8.73

8.72

8.70

9.24

9.16

9.11

9.05

9.00

8.95

8.91

8.88

8.68

8.66

8.65

8.64

8.64

11.54

11.44

11.33

11.19

11.04

10.97

10.90

12.58

12.41

12.26

12.10

12.00

11.88

11.72

11.67

10.80

10.70

10.59

10.48

10.37

9.89

9.85

9.81

9.77

9.73

9.68

9.66

10.30

10.21

10.14

10.06

9.99

9.95

9.61*

9.57

9.54*

9.52

9.47*

9.43

9.38*

119

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

85

90

95

65

70

75

80

50

55

60

30

35

40

45

15

20

25

Temp (°C)

0

5

10

20.1.8 WTW buffers

1.99

1.99

1.98

1.98

1.98

1.98

1.98

pH of buffer solutions

2.03

4.01

2.02

2.01

4.01

4.00

2.00

2.00

2.00

4.00

4.00

4.01

1.99

4.01

4.02

4.03

4.04

4.06

4.08

4.10

4.13

4.16

4.19

4.22

4.26

4.30

4.35

20.1.9 JIS Z 8802 buffers

55

60

70

80

90

95

38

40

45

50

20

25

30

35

Temp (°C)

0

5

10

15 pH of buffer solutions

1.666

4.003

1.668

1.670

1.672

3.999

3.998

3.999

1.675

1.679

1.683

1.688

1.691

1.694

1.700

1.707

4.002

4.008

4.015

4.024

4.030

4.035

4.047

4.060

1.715

1.723

1.743

1.766

1.792

1.806

4.075

4.091

4.126

4.164

4.205

4.227

6.99

6.98

6.97

6.97

6.97

6.98

6.98

7.12

7.09

7.06

7.04

7.02

7.00

6.99

7.00

7.02

7.04

7.06

7.09

7.12

6.984

6.951

6.923

6.900

6.881

6.865

6.853

6.844

6.840

6.838

6.834

6.833

6.834

6.836

6.845

6.859

6.877

6.886

9.464

9.395

9.332

9.276

9.225

9.180

9.139

9.102

9.081

9.068

9.038

9.011

8.985

8.962

8.921

8.885

8.850

8.833

10.65

10.52

10.39

10.26

10.13

10.00

9.87

9.74

9.61

9.48

9.35

120

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

Transmitter M300

20.2 Dual membrane pH electrode buffers

20.2.1 Mettler-pH/pNa buffers (Na+ 3.9M)

Temp (°C)

0

25

30

35

5

10

15

20

40

45

50 pH of buffer solutions

1.98

3.99

1.98

1.99

1.99

1.99

2.00

2.00

2.01

2.01

2.02

2.02

4.05

4.07

4.09

3.99

3.99

3.99

4.00

4.01

4.02

4.04

7.01

7.00

7.00

6.99

7.00

7.00

7.01

7.01

7.02

7.03

7.04

9.51

9.43

9.36

9.30

9.25

9.21

9.18

9.15

9.12

9.11

9.10

121

© 01 / 2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland

Printed in Switzerland

Transmitter Multiparameter M300

52 121 389

METTLER TOLEDO Market Organizations

Sales and Service:

Australia

Mettler-Toledo Limited

220 Turner Street

Port Melbourne, VIC 3207

Australia

Phone +61 1300 659 761 e-mail [email protected]

Austria

Mettler-Toledo Ges.m.b.H.

Laxenburger Str. 252/2

AT-1230 Wien

Phone +43 1 607 4356 e-mail [email protected]

Brazil

Mettler-Toledo Ind. e Com. Ltda.

Avenida Tamboré, 418

Tamboré

BR-06460-000 Barueri/SP

Phone +55 11 4166 7400 e-mail [email protected]

[email protected]

Canada

Mettler-Toledo Inc.

2915 Argentia Rd #6

CA-ON L5N 8G6 Mississauga

Phone +1 800 638 8537 e-mail [email protected]

China

Mettler-Toledo International Trading

(Shanghai) Co. Ltd.

589 Gui Ping Road

Cao He Jing

CN-200233 Shanghai

Phone +86 21 64 85 04 35 e-mail [email protected]

Croatia

Mettler-Toledo d.o.o.

Mandlova 3

HR-10000 Zagreb

Phone +385 1 292 06 33 e-mail [email protected]

Czech Republic

Mettler-Toledo s.r.o.

Trebohosticka 2283/2

CZ-100 00 Praha 10

Phone +420 2 72 123 150 e-mail [email protected]

Denmark

Mettler-Toledo A/S

Naverland 8

DK-2600 Glostrup

Phone +45 43 27 08 00 e-mail [email protected]

France

Mettler-Toledo

Analyse Industrielle S.A.S.

30, Boulevard de Douaumont

FR-75017 Paris

Phone +33 1 47 37 06 00 e-mail [email protected]

Germany

Mettler-Toledo GmbH

Prozeßanalytik

Ockerweg 3

DE-35396 Gießen

Phone +49 641 507 444 e-mail [email protected]

Great Britain

Mettler-Toledo LTD

64 Boston Road, Beaumont Leys

GB-Leicester LE4 1AW

Phone +44 116 235 7070 e-mail [email protected]

Hungary

Mettler-Toledo Kereskedelmi KFT

Teve u. 41

HU-1139 Budapest

Phone +36 1 288 40 40 e-mail [email protected]

India

Mettler-Toledo India Private Limited

Amar Hill, Saki Vihar Road

Powai

IN-400 072 Mumbai

Phone +91 22 2857 0808 e-mail [email protected]

Indonesia

PT. Mettler-Toledo Indonesia

GRHA PERSADA 3rd Floor

Jl. KH. Noer Ali No.3A,

Kayuringin Jaya

Kalimalang, Bekasi 17144, ID

Phone +62 21 294 53919 e-mail [email protected]

Italy

Mettler-Toledo S.p.A.

Via Vialba 42

IT-20026 Novate Milanese

Phone +39 02 333 321 e-mail [email protected]

Japan

Mettler-Toledo K.K.

Process Division

6F Ikenohata Nisshoku Bldg.

2-9-7, Ikenohata

Taito-ku

JP-110-0008 Tokyo

Phone +81 3 5815 5606 e-mail [email protected]

Malaysia

Mettler-Toledo (M) Sdn Bhd

Bangunan Electroscon Holding, U 1-01

Lot 8 Jalan Astaka U8/84

Seksyen U8, Bukit Jelutong

MY-40150 Shah Alam Selangor

Phone +60 3 78 44 58 88 e-mail

[email protected]

Mexico

Mettler-Toledo S.A. de C.V.

Ejército Nacional #340

Polanco V Sección

C.P. 11560

MX-México D.F.

Phone +52 55 1946 0900 e-mail [email protected]

Norway

Mettler-Toledo AS

Ulvenveien 92B

NO-0581 Oslo Norway

Phone +47 22 30 44 90 e-mail [email protected]

Poland

Mettler-Toledo (Poland) Sp.z.o.o.

ul. Poleczki 21

PL-02-822 Warszawa

Phone +48 22 545 06 80 e-mail [email protected]

Russia

Mettler-Toledo Vostok ZAO

Sretenskij Bulvar 6/1

Office 6

RU-101000 Moscow

Phone +7 495 621 56 66 e-mail [email protected]

Singapore

Mettler-Toledo (S) Pte. Ltd.

Block 28

Ayer Rajah Crescent #05-01

SG-139959 Singapore

Phone +65 6890 00 11 e-mail [email protected]

Slovakia

Mettler-Toledo s.r.o.

Hattalova 12/A

SK-83103 Bratislava

Phone +421 2 4444 12 20-2 e-mail [email protected]

Slovenia

Mettler-Toledo d.o.o.

Pot heroja Trtnika 26

SI-1261 Ljubljana-Dobrunje

Phone +386 1 530 80 50 e-mail [email protected]

South Korea

Mettler-Toledo (Korea) Ltd.

1&4F, Yeil Building 21

Yangjaecheon- ro 19-gil

SeoCho-Gu

Seoul 06753 Korea

Phone +82 2 3498 3500 e-mail [email protected]

Spain

Mettler-Toledo S.A.E.

C/Miguel Hernández, 69-71

ES-08908 L’Hospitalet de Llobregat

(Barcelona)

Phone +34 902 32 00 23 e-mail [email protected]

Sweden

Mettler-Toledo AB

Virkesvägen 10

Box 92161

SE-12008 Stockholm

Phone +46 8 702 50 00 e-mail [email protected]

Switzerland

Mettler-Toledo (Schweiz) GmbH

Im Langacher, Postfach

CH-8606 Greifensee

Phone +41 44 944 47 60 e-mail [email protected]

Thailand

Mettler-Toledo (Thailand) Ltd.

272 Soi Soonvijai 4

Rama 9 Rd., Bangkapi

Huay Kwang

TH-10320 Bangkok

Phone +66 2 723 03 00 e-mail

[email protected]

Turkey

Mettler-Toledo Türkiye

Haluk Türksoy Sokak No: 6 Zemin ve 1.

Bodrum Kat 34662 Üsküdar-Istanbul, TR

Phone +90 216 400 20 20 e-mail [email protected]

USA

METTLER TOLEDO

Process Analytics

900 Middlesex Turnpike, Bld. 8

Billerica, MA 01821, USA

Phone

Freephone +1 800 352 8763 e-mail

+1 781 301 8800 [email protected]

Vietnam

Mettler-Toledo (Vietnam) LLC

29A Hoang Hoa Tham Street, Ward 6

Binh Thanh District

Ho Chi Minh City, Vietnam

Phone +84 8 35515924 e-mail

[email protected]

ISO

9001 certified

ISO

14001 certified

Management System certified according to

ISO 9001 / ISO 14001

N315

Subject to technical changes.

© Mettler-Toledo GmbH, Process Analytics

01/2016 Printed in Switzerland. 51 121 389