Midmark International (Whip) Delivery offers precise control over fluid delivery, allowing for accurate and efficient fluid administration. Its advanced design enables a wide range of flow rates for versatile use in various medical applications. The device's user-friendly interface and intuitive controls enhance ease of operation, making it ideal for a variety of clinical settings. Trust Midmark International (Whip) Delivery for reliable and precise fluid delivery, ensuring optimal patient care.
advertisement
Midmark
®
Dental Light, Console Mounted on a Knight
Chair Setup Guide
waRning
Equipment is not suitable for use in the presence of a flammable anesthetic mixture.
Stop Ring & Pin Installation
nOTE: This setup guide applies to Halogen and LED-style lights.
Halogen LED
Applies to Models:
153937 (-001 & -002)
153938 (-001 & -002)
Set
Screw
Stop
Pin
Cover
Ring
Flex arm
Shaft
Rotation
Stop Ring
Curved arm assembly
Step 1:
A) Position tab of rotation stop ring in line with flex arm, facing lighthead as shown.
B) Insert flex arm shaft into curved arm assembly & secure with set screw.
C) Lift cover ring & install stop pin in curved arm assembly.
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com
English-1 Style G 003-1121-00 Rev. TA 1 (4/ 15 /19)
Light Arm Setup
(with asepsis 21 console)
Step 1: Remove both side covers from console
& loosen two set screws.
Step 4: Install bearing washer & trim ring onto curved arm.
Step 3: Install pivot bearing & insert light post assembly into center hole of console.
note
The lower bearing will rest on top of mounting screws of the accessory mounting plate.
Pivot
Bearing
Lower
Bearing
Step 2: Install lower bearing, then tighten set screws.
Set Screw accessory
Mounting Plate
Curved arm wire Harness
Harness
Connection
Trim
Ring
Step 5: Connect curved arm wire harness to jumper harness as shown.
Bearing washer
Jumper
Harness
Step 6: Insert curved arm into center hole of console, feeding wire harness down thru console mounting bracket.
Step 7: Route wire harness up thru bottom of console and connect as shown.
Power
Supply
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com
English-2 Style G
Chair
Light
Harness
Step 8: Connect chair light harness to power supply.
Step 9: Verify proper operation of light and light arm.
Tension adjustments
neck Cover
(Long Neck)
To adjust tension at Light pivot joint...
A) Remove screw cover as shown.
B) Loosen set screw.
C) Loosen or tighten light pivot adjustment screw.
D) Tighten set screw.
To adjust tension at arm / Yoke pivot joints...
A) Remove neck cover as shown (see Note).
B) Loosen or tighten appropriate screws to adjust tension.
Note: Some halogen models have a short neck on the lighthead. It is not necessary to remove the neck cover on these models.
Light Pivot adjustment Screw
Short neck arm Pivot adjustment Screw
Yoke Pivot adjustment Screw arm Pivot adjustment Screws
Set Screw
Yoke Pivot adjustment Screws
To adjust Flex arm tension...
A) Remove end cap and top cover as shown.
B) Loosen or tighten adjustment nut as desired.
adjustment nut
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com
English-3 Style G