AMS AS89010 Sensor Interface User Guide
The AMS AS73210 is a small PCB for general color measurement and control applications, with a high bandwidth of light energy and handheld applications. It includes a True Color sensor AS73210 based on the international CIE 1931 standard, a special analog-to-digital converter with a high dynamic range, an EEPROM for sensor data, an integrated temperature sensor, and 2 LDO micro power regulators.
Iklan
Iklan
high performance needs great design.
Coverpage: AS73210-AS89010-AB4 SET DK User Guide
Please be patient while we transfer this adapted former
MAZeT document to the latest ams design. ams.com db13386e_V2.01_certified_2017-07-06
User Guide
AS73210-AS89010-AB4 SET DK
Development Kit
Order No.: 220800004
Status: certified
INTRODUCTION
The family of AS73210-AS89010-AB4 based test systems of ams Sensors Germany, their housing, adapters and further information are described in this document.
The Test SETs include the OEM AS73210-AS89010-AB4 sensor board [5], optional (Development Kit) fitted into a metallic housing with optical interface, a standard I²C-to-USB converter as well as Windows PC test software.
The test software enables to control the converter, calibrate the sensor and allows data logging options including a print out of results. Please note, the development kit was designed to be a test system for system and application tests.
Please note, the development kit was designed to be a test system for using only in system and application tests.
Figure 1: Block Diagram AS73210-AS89010-AB4
The OEM sensor board AS73210-AS89010-AB4 is a small PCB for general color measurement and control applications, with a high bandwidth of light energy and handheld applications. The sensor board includes a
True Color sensor AS73210 [2] based on the international CIE 1931 standard, a special analog-to-digital converter AS89010 [1] with a high dynamic range (1 : 1,000,000), an EEPROM for sensor data, an integrated temperature sensor, 2 LDO micro power regulators (LT1761ES5-3, Linear Technology) to manage the analog and digital supply voltage. I²C is used for external communication, configuration of the sensor, readout of the sensor data as well as writing and reading of the memory.
⚖ Production data information is current as of publication date. Products conform to specifications per the terms of ams Sensors Germany GmbH. The information in this document is subject of change without notice, please confirm that this is the latest version. Please check with a ams Sensors Germany sales representative for availability and further information. Full legal notices can be found on the final page.
AS73210-AS89010-AB4 SET DK
V2.01
The OEM sensor board AS73210-AS89010-AB4 [5] uses a standard I²C-to-USB converter for initial testing reasons via USB and PC software. This converter IOW24-DG [4] is connected to the sensor board via cable and controls uncritical functions (not synchronized) of the sensor board via PC.
The standard software is only valid for Windows based systems (Minimum Windows 7). The I²C to USB converter [4,6] also supports LINUX and iOS based systems.
The description of the sensor, converter, OEM hardware and I²C communication is implemented in the reference documents [1] [2] [3] and [5]. For all details of the I²C converter, see [4].
The following chapters describe the development kit and especially the software setup and usage for a
Windows PC. Please make sure to install all libraries and test software before you connect the development kit on PC via USB. Otherwise, the test software starts in a demonstration mode and does not make real measurements. ams.com db13386e_V2.01_certified_2017-07-06 Page 2 of 30
AS73210-AS89010-AB4 SET DK
V2.01
ams.com db13386e_V2.01_certified_2017-07-06 Page 3 of 30
AS73210-AS89010-AB4 SET DK
V2.01
1 SCOPE OF DELIVERY
The AS73210-AS89010-AB4 product family includes alternative delivery forms with the following standard elements:
▪ Sensor board AS73210-AS89010-AB4 (housed or OEM PCB)
▪ Setup for PC test are including setup for MATLAB Compiler Runtime, not for OEM version
▪ I²C cable with USB converter (device from [4]) , not for OEM and/or optional
▪ Optical cover for realize angle of incidence, not for OEM and/or optional
Defined in [6] alternative IO-Warriors and libraries are available for customers wanting to integrate the kit into own software environments. For the scope of delivery of alternatives see the following table:
Table 1: Scope of delivery for AS73210-AS89010-AB4 product family (form of deliveries may be different)
Name AS73210-AS89010-AB4 SET DK
JENCOLOR ® Development Kit with USB-Interface
AS73210-AS89010-AB4
JENCOLOR ® OEM Sensor Board with I²C-Interface
220800004
O
I²C / USB via dongle
O
Article
Sensor AS73210 + AS89010
Interface
PC Windows Test software
Housing and adapter
USB cable
JCDK-Aperture-13.5
Optical Cover
O / O
O
O optional
220800002
O
I²C
---
O / -
--- optional
O
The housed test system consists from a basic case which will be adapted customized and/or application specific by various parts and optical components (apertures).
The solid basic case was made of anodized aluminum, which protects the sensor board AS73210-AS89010-
AB4 against mechanical stress as well as ESD and EMF. The sensor AS73210 is positioned in the basic case coaxial with a 3/8”-24 UNF internal thread and ready for the application of different optical accessories. ams.com db13386e_V2.01_certified_2017-07-06 Page 4 of 30
AS73210-AS89010-AB4 SET DK
V2.01
Table 2: Basic case and adapters for the DK AS73210-AS89010-AB4 1
JCDK-Basic case JCDK-Aperture-13.5 JCDK-Aperture-13.5 with IR-cut-off filter
JCDK-1/4“-
Adapter
JCDK-SMA-
Adapter
There are difference adapters “Aperture” available to adapt for the sensor in the package the angle of incidence and/or to fix the sensor system.
The JCDK-Aperture-135 is designed to limit the angle of incidence on the AS73210 to ±10°. It is an optional aperture of the JCDK assortment and characterized with a single groove outside. The outside diameter of
25mm makes it compatible with JCDK-1/4”-Adapter and opto-mechanical system parts (e.g. Linos). In front of the aperture is space for mounting a filter glass with outside diameter up to 22.2 mm and thickness up to mm.
JCDK-1/4“-Adapter is used for mounting the JENCOLOR ® DKs on a camera tripod or equivalent holders. On the bottom side there is a centered 1/4”-20 UNC internal thread and 20mm below the optical axis there are two M4 internal threads for mounting the JCDK-1/4“-Adapter on a flat surface. For IR sensible applications order the JCDK aperture 13.5 with inserted IR blocking. For using the adapter with an alternative integrated filter please ask our sales team.
Via the JCDK-SMA-Adapter a fiber optic patch cable with F-SMA 905 plug can be connected to the
JENCOLOR ® Development Kit. The distance between fiber end and sensor is calculated to achieve an angle of incidence of at most 10° with a fiber core diameter up to 600µm. Minimum NA for illumination of the whole sensor area is 0.22 and to reduce the effect of stray light inside the adapter and the case NA should not exceed 0.39.
1
For other adapters please contact our sales team
. ams.com db13386e_V2.01_certified_2017-07-06 Page 5 of 30
AS73210-AS89010-AB4 SET DK
V2.01
2 SYSTEM REQUIREMENTS
For start-up procedures, the following system resources are required:
▪ PC Pentium 1 GHz or higher
▪ 512 MB RAM
▪ 1x free USB (2.0) port
▪ 80 MB free hard drive memory (+ approx. 1.3 GB for MATLAB Compiler Runtime)
▪ Microsoft Windows 7 and above
▪ (optional 2 for import/export functions) Microsoft EXCEL
▪ Internet Connection and administration right for runtime and software installation
3 SPECIFICATION OF AS73210-AS89010-AB4 FAMILY
Table 3: Specification of (SET) (DK) AS73210-AS89010-AB4
Spectral range
380nm … 700nm
Measuring Method
Measuring values
Based on XYZ CIE 1931 standard
(user calibration necessary) 3
ADC digit; ADC current;
XYZ;
Yxy; Yxy SDCM; Yuv
Gain (reference currents)
Integration time
Digital electronic resolution
Selectable 20nA …5120nA (5 stages)
Selectable 1ms … 1sec (11 stages)
16bit ADC output by using divider for 20 internal bits
LSB min lux
FSR min lux
FSR max lux
Dynamic range
Power supply
PC interface
Weight
Operating conditions
0.00005
3.31
13,576.43
1 – 1,000,000
USB powered
USB 2.0 full speed
60 g plus adapter ca. 5…10g
Temperature 0 ... 85 °C
Humidity 85% relative humidity at 35°C
2
E.g. recommend to use some special program functions like write protocol and others
3 For more details for sensor response, sensitiveness and other parameters see the AS73210 datasheet ams.com db13386e_V2.01_certified_2017-07-06 Page 6 of 30
AS73210-AS89010-AB4 SET DK
V2.01
4 START-UP
4.1 Software/Driver Installation
This section of the document describes how to install and use the AS73210-AS89010-AB4 PC-software.
During process installation and work administration permission are needed. Note: The SETUP must be run with the special option “run with administration rights”.
The installation process also requires a working internet connection to connect to MATLAB-specific system requirements. Please check your system configuration in Windows System Control for the Internet connection and right settings.
Before you start the SETUP file for the test software check whether the actual MATLAB Compiler Runtime
(MCR) 8.2 version (standard installation by MCR_R2013b_win32_installer.exe) is installed on your PC 4 . If
MATLAB Compiler Runtime library was not installed then follow the special instructions in the setup process for MATLAB.
To start the installation of the AS73210-AS89010-AB4 test software simply execute the setup.exe file from the USB stick or other memory which you got and all required files are installed automatically. The setup opens in a window which guides the user through the installation step by step.
Figure 2: Setup - Welcome screen Setup
The setup software checks the installed MCR version. If the version is incorrect the setup attempts to download the correct MCR via internet. If your internet connection requires a proxy server, configure the server using the ‘Connection Settings’ button.
Figure 3: Setup - Connection settings
4
The setup routine always check internally whether Setup Matlab was defined by the System variables $PROGRAMFILES. Normally that is the path C:\Program Files (x86)\MATLAB\MATLAB Compiler Runtime. ams.com db13386e_V2.01_certified_2017-07-06 Page 7 of 30
AS73210-AS89010-AB4 SET DK
V2.01
Click ‘Next’ to specify installation options, set the installation path and add a shortcut to the desktop. The installation folder will include the program and data files in several directories.
Figure 4: Setup - Destination Folder
After clicking ‘Next’ setup will check the installed MCR version. If the correct version was installed just click
‘Next’ at the following screen. Otherwise, setup attempts to download the correct version via internet and installs it.
Figure 5: Setup - MATLAB Compiler Runtime version already installed. ams.com db13386e_V2.01_certified_2017-07-06 Page 8 of 30
AS73210-AS89010-AB4 SET DK
V2.01
Figure 6: Setup - Download required Compiler Runtime and confirm MATLAB License Agreement
In the next screen, there is the option to confirm the MathWorks License Agreement.
Figure 7: Setup - Start installation
The software will be installed into the shown destination folders after confirming via ‘Install’ button. ams.com db13386e_V2.01_certified_2017-07-06 Page 9 of 30
AS73210-AS89010-AB4 SET DK
V2.01
Figure 8: Setup - Installing
The installation is completed after confirming via 'Finish' button.
Figure 9: Setup - Installation finished
4.2 Remove
To uninstall the software use the Windows system setup “Add or remove programs”. Choose the programs to uninstall them in the following dialog.
Figure 10: Setup - Remove program
Press ‘Uninstall’ to delete the complete software package. You may be required to manually remove some files remaining in the installation of the software. ams.com
Attention: Create a backup of your own created files that are located within the installation folder. db13386e_V2.01_certified_2017-07-06 Page 10 of 30
AS73210-AS89010-AB4 SET DK
V2.01
5 SOFTWARE DESCRIPTION
5.1 Software start
Use the standard to start the AS73210-AS89010-AB4 software by the installed icon MTCS_INT_AB4 in the program menu. In the main user interface the specification for the measurements e.g. integration time, reference currents, divider and offset must be selected before the measurement process will be started
Software AS73210-AS89010-AB4\application” (see chapter 5.8).
Figure 11: Standard properties of the installed AS73210-AS89010-AB4 software
Another variant to start the software DK AS73210-AS89010-AB4 with specified parameters is to expand the start command in the property by a pre-defined configuration file. See the following example for a complete text in the target of the MTCS_INT_AB4 property to start the software:
"……. MTCS_INT_AB4.exe" "command example AS89010-EB1.csv"
5.2 Main User Interface Elements
After a successful installation and running the software the user interface always starts in Expert Mode
parameters for measurements (Configuration), to select form of outputs (Diagram), and to control the next program steps (Measurement, Calibration). ams.com db13386e_V2.01_certified_2017-07-06 Page 11 of 30
AS73210-AS89010-AB4 SET DK
V2.01
Figure 12: Main User interface elements in Expert Mode
Configuration: The most important parameters can be changed direct by the user. It is possible to initialize
starts a measurement with the selected initialization. Based on the selected mode the measurement will be performed sequentially by command after each click and/or in continuous mode. By selecting the parameters of the converter like Integration Time, Reference Current, Divider and Offset parts of the gain of the sensor system can be directly controlled by the user. If an initialization via a *.csv-file (see
chapter 5.8) was selected then Configuration shows the actual selected parameters of *.csv. For more
details and parameters of Configuration please see chapter 5.5 and/or the datasheet of AS89010 [1].
is performed, the status of calibration is invalid and the button is highlighted in red. After completing a calibration the status is valid and the button is highlighted in green. A change in the configuration parameters changes the status of the calibration back to invalid (red) if the parameters differ from the setting used for the calibration.
Diagram: The representation of the measured data and a table for the data output of the measurements can be enabled / disabled and adjusted.
Measure / Reference: A (reference) measurement (see chapter 5.3 and/or 5.4) is started with the
selected parameters.
The Clear button deletes all measurement values and closes the Color space window. It does not delete any register settings 5 . The following example shows the color space Yu’v’. (Figure 15)
5 Please note that register settings only could be changed by new writing or after a hardware reset to default via USB disconnection/connection. ams.com db13386e_V2.01_certified_2017-07-06 Page 12 of 30
AS73210-AS89010-AB4 SET DK
V2.01
Status: The status window shows important steps performed by the program and I²C status is logged (see
In the Menu Bar different functions for the initialization of the program, parameters and/or data input / output are listed.
Load/Save: In Load an initialization file for the gain parameters can be read (‘Load Config’, see chapter
calibration and measured data’s of a last session. In Save the entry ‘Save Config and measure data’ saves all data of the actual session in a *.dat file and ‘Export…’ writes all configuration data’s and the measured results in a *.xls
6 or *.csv file.
measurements after a calibration has been performed.
5.3 Easy Mode
The Easy Mode is designed to be as simple as possible using only limited functions. It can be set manually in menu bar as shown in the following picture in case the expert mode is active.
Figure 13: Menu bar - Easy mode
Figure 14: Main User interface elements in Easy Mode
To configure the device in the Easy Mode you can load a predefined configuration *.csv file or a full session file (*.dat) including configuration, calibration and measurements. To do this use the Load menu.
After pressing the Measure button a color chart appears with the selected diagram which shows the measured values of the sensors. Using a command based configuration file, clicking generates one new measurement value (in command mode) or continues measurements (in continues mode) until the stop button is pressed. The result(s) of the measurement(s) are shown in the selected diagram color space window as a figure, or, if activated, in a data table under the figure.
After starting the program this data table is not automatically activated to increase the speed of the measurements. If Data table is activated, please note the final result of the measurement is always displayed in the first line.
6 Excel installation on PC is necessary ams.com db13386e_V2.01_certified_2017-07-06 Page 13 of 30
AS73210-AS89010-AB4 SET DK
V2.01
The number of measurements is always shown in row one - name.
Figure 15: Measurement values – with selected Yu'v' diagram and selected Data table
5.4 Expert Mode
The Expert Mode offers further options like a reference measurement, sensor calibration, and a detailed
I 2 C status report. The Expert Mode can be activated similar to the Easy Mode via menu button Mode. The
user interface changes automatically after choosing this option as shown in the Figure 16.
[Direction In: Only to use in combination with the AS89010 test board. Otherwise use it as “Direction In”. ]
Calibrate:The button Calibrate helps you to perform a sensor/application calibration to optimize the
overall accuracy based on the CIE 1931 standard. Read section 5.6 for further information. If the
Calibrate button is highlighted in red, it indicates that no calibration file is loaded and that the sensor only shows exact ADC values.
After the calibration has been performed and the Calibration button turns green all selected color points shown in the color space windows are shown exactly according to the calibration.
Status: The Status field provides information about the run time processes. This can be for example the
ADC value, readings or settings of the configuration on the sensor board via I 2 C or a specific calibration status.
Reference: Pressing the Reference button while color spaces are selected in the Diagram dropdown performs a comparison of the reference and measured values in the selected color space. Subsequently, pressing the Measure button allows to view the delta values between the reference and the measured values. To generate a new reference value, just press the Reference button again and the user interface will be updated with the new reference value. ams.com db13386e_V2.01_certified_2017-07-06 Page 14 of 30
AS73210-AS89010-AB4 SET DK
V2.01
Figure 16: Expert mode - User interface ams.com db13386e_V2.01_certified_2017-07-06 Page 15 of 30
AS73210-AS89010-AB4 SET DK
V2.01
Figure 17: Reference measurement – with selected Yxy diagram and selected data table
5.5 Basics for Measurements
Each ADC value (Analog-Digital Converted value) is calculated to a photocurrent (“photocurrent_nA”) under consideration of dark current (“zero”), reference current (“refCurrent_nA”) and integration time (“nClck”).
These steps are performed before this value is used for any calibration or colorimetric functions. The following virtual code is used for each sensor channel value - whereas the ADC is the 16bit read out value:
% Correction ofbit shift (register CREGH)
% divider = 1 (off), 2, 4, 8 or 16
Adc = Adc * divider;
% Calcualtion of the maximum number of clocks nClck
% intTime_ms is the integration time in milliseconds (register CREGL) bit = 10 + log2(intTime_ms); nClck = 2^bit;
% Check of saturation sat = nClck - 1; if sat > 65535
sat = 65535; end if Adc >= sat
% Error! Sensor channel is saturated! ams.com db13386e_V2.01_certified_2017-07-06 Page 16 of 30
AS73210-AS89010-AB4 SET DK
V2.01
else
% Measurement is ok. end
% Correction of zero offset (register OPTREG)
% zero = 0 (disabled), 15, 31 or 63
Adc = Adc - zero;
% Calculation of photo current in nano ampere photocurrent_nA = Adc * refCurrent_nA / nClck;
A detailed description of the sensor signal calculation is included in the datasheets of JENCOLOR ® key components, white papers or application notes. Please ask our sales team for further information.
To perform absolute color measurements or for further evaluation of the results in any color space, the sensors system DK AS73210-AS89010-AB4 must be calibrated to the specific application or test setup.
Please ensure that all continuous processes, measurements and the calibration procedures are performed based on same conditions. These aspects refer to the DUT (device under test), distances between DUT and sensor as well as all initialization and environment conditions. The sensor, selected illumination and target are used in a closed system for calibration and measurement procedures. A new calibration is required if you change any conditions inside this closed system.
5.6 Calibration
The calibration option is used to create an application and sensor specific calibration via calibration targets measured by a reference device like a spectrometer. The standard calibration simply uses an identity matrix which approximates the color space values. Therefore it is recommended to calibrate the sensor/application to get accurate results and coordinates in any color spaces. This section describes how to calibrate the device.
First set the observer with which the reference values are measured (see colorimetric settings of your reference device alike a spectrometer or others).
Figure 18: Calibration - Observer
Then choose whether the calibration matrix should be inserted manually or to measure it.
Figure 19: Calibration - Targets measure or manually - ams.com db13386e_V2.01_certified_2017-07-06 Page 17 of 30
AS73210-AS89010-AB4 SET DK
V2.01
After choosing Import the calibration matrix and offset values must be inserted e.g. from an Excel sheet like shown in the following figure. Just select the cells for import and press the button Import from xls.
Figure 20: Calibration - Manual, insert matrix
In the next step the offset values for the sensor must be imported.
Figure 21: Calibration - ADC offset values
After choosing the Measure option, it is required to enter the amount of calibration targets those want to be measured. The minimal amount of targets is the number of sensor channels. The DK AS73210-AS89010-
AB4 uses at least 3 channels.
Next, the software requests to cover the sensor for an offset measurement. After this following the software performs measurements for each target and guides the user through the process step by step.
Figure 22: Calibration - Measurement steps Dark measurement and measurements of targets -
In the final step, the software requests to insert XYZ values measured from the reference device. It is possible to save the results of the calibration process into a special owner configuration file.
If any support during the sensor calibration is needed, please contact our sales team. ams.com db13386e_V2.01_certified_2017-07-06 Page 18 of 30
AS73210-AS89010-AB4 SET DK
V2.01
Figure 23: Calibration - Input XYZ reference values -
A calibration is valid for the used system and must not be re-calibrated until not relevant system configurations or parameters for measurement were changed (e.g. integration time). After changing the configuration file the software always tests the specified parameters if a re-calibration is necessary and shows the message "No calibration loaded” or “Calibrated with different settings”. In this case the calibration must be performed again to achieve exact results. Please note, the software only tests known parameters of the test software and cannot check if any external conditions in the test system were changed. Ask our sales team for support during the calibration process or if you will not get the specified results after calibration.
5.7 Load/Save menu
The Load and Save menus provide the option to load and save the configuration, calibration and measurement values in a file with the extension *.dat (‘Load config and measure data’; ‘Save config and measure data’). To simply load a configuration of the sensor use the Load config menu entry. This allows
you to load a *.csv (*.csv – comma separated values) configuration file (see chapter 5.8).
The Save menu also offers an extra feature to export measurement values to *.csv files or the actual sensor configuration and measurement values to an Excel file (*.xls) for further usage. This data can be used for comparison or for statistical calculations. Note: The export to Excel is only available if Excel is properly installed on your local PC.
Figure 24: Menu - Load and Save
5.8 Configuration Files (*.csv)
The Configuration Files (*.csv – comma separated values) contain all important information about the measurement and configuration behavior. The files are used for application-specific adjustment of the test board AS73210-AS89010-AB4 and/or compatible boards. In the installation path of the test software you find several examples for a configuration file like “command example.csv” or “command example AS89010-
EB1.csv” 7 . Processing these *.csv-files is possible with a text editor, Open Office Software or MS Excel.
Please note, however compliance with the syntax and the extension *.csv when saving. Other-wise, errors
7 This file is an example for a software adaptation for an alternative and AS73210-AS89010-AB4 compatible test board AS73210-
AS89010-EB1. This board represents an evaluation kit for ams Sensors Germany’s ADC AS89010 and I²C in interface. ams.com db13386e_V2.01_certified_2017-07-06 Page 19 of 30
AS73210-AS89010-AB4 SET DK
V2.01
will occur during software operation. Table 1 indicates all relevant syntax, columns and register commands.
All grey highlighted columns are comments or user information with no conditions for syntax except for the use of the separation of characters like “,” or “;”. In case of using JENCOLOR ® sensors a changing of the line with the keyword “ChannelOrder” is not necessary because the filter set was pre-defined. On the other side by using alternative sensor boards the ChannelOrder must be adapted based on the any special configuration. Check the installation or support files like readme.txt or datasheet for more details.
To use a configuration file, select the file and load it by mouse click. Then the parameters inside the files specified will be used. ams.com db13386e_V2.01_certified_2017-07-06 Page 20 of 30
AS73210-AS89010-AB4 SET DK
V2.01
Table 4: Command example.csv - in includes not a scenario for measurement but different alternative examples to show structure and coding
STATE INFO
REGISTER
INFO DATA COMMENT R/W I 2 C
ADDRESS
REGISTER DATA
ChannelOrder specifies the number and reference of the filters .e.g.
JENCOLOR ® sensor
1 3
Init
Init
W
W
E8
E8
0
0
2
0
Init
Init
Init
Init
Init
Init
Init
Init
OSR
OSR
CREGL
CREGL
CREGL
CREGL
CREGL
CREGH
CREGH
OPTREG
Set config mode Access to config registers
No power down mode
Wake up MCDC (wait 500 µs before measurement)
R=20 nA T=1 ms R=reference current
T=integration time
R=20 nA T=1024 ms R=reference current
T=integration time
R=5.12 µA T=1 ms R=reference current
T=integration time
R=5.12 µA
T=1024 ms
R=reference current
T=integration time
R=1.28 µA
T=128 ms
CMD Div=off
R=reference current
T=integration time
Command mode and digital divider off
CMD Div=2
Zero=63
Command mode and digital divider 2
Set offset of 63 digits
(recommend)
W
W
W
W
W
W
W
W
E8
E8
E8
E8
E8
E8
E8
E8
6
6
6
6
6
7
7
8
128
138
192
202
183
8
9
3 ams.com db13386e_V2.01_certified_2017-07-06
2
Page 21 of 30
AS73210-AS89010-AB4 SET DK
V2.01
STATE INFO
REGISTER
Init OSR
Init OSR
INFO DATA
Set address to register
Read 11 byte
COMMENT
Set address point to register
Measure
Measure
Measure
OSR
OUT0
OUT0
Start measure
Set address to register
Read 8 byte
Read all config registers (if access = R*)
Begin single measurement
Set address to first output register
Read OUT0 to OUT3
R/W
W
I 2 C
ADDRESS
E8
REGISTER DATA
0
R E9 11
W
W
R
E8
E8
E9
0
0
131
8 ams.com db13386e_V2.01_certified_2017-07-06 Page 22 of 30
AS73210-AS89010-AB4 SET DK
V2.01
All commands and data refer to the I²C communication and register settings of the Sensor signal ASIC AS89010. For more details please check the description in the datasheet of the AS89010 [1].
Description of *.csv columns:
“State” All rows containing “Init” are send to the AS73210-AS89010-AB4 to initialize the AS89010 after choosing a configuration .
Important: Do only use the commands “ChannelOrder”, “Init” and “Measure” (case sensitive)! Do not mix “Init” and “Measure” rows!
“Info Register”,
“Info Data”, Columns are not relevant for a functional use.
“Comment”
“Read/Write” Insert “W” for write I 2 C data or “R” for read I 2 C data.
“I 2 C Address” Insert “E8” for write and “E9” for read I 2 C data.
“Register” For write I 2 C: Insert the register address which should be written to. Important: Only insert decimal values 0...255!
For read I 2 C: Leave it empty.
“Data” For write I 2 C: Insert the data which should be written. Important: Only insert decimal values 0...255!
For read I
2
C: Insert the amount of bytes which should be read.
Attention:
The JENCOLOR ® AS73210-AS89010-AB4 Evaluation Software is designed like as a simple terminal program. This means that each write and read command
(each row of configuration file) is written or read via I 2 C without checking the validation of data. For detailed description of all registers check the datasheet of
AS89010 [1]. The JENCOLOR ® AS73210-AS89010-AB4 Evaluation Software can only be used in Command and Continuous mode (compare register CREGH of
AS89010).
After software installation in the software directory “application” some samples of *.csv files show several examples to set the system, write/read registers and measure in different modes. ams.com db13386e_V2.01_certified_2017-07-06 Page 23 of 30
AS73210-AS89010-AB4 SET DK
V2.01
6 QUICK START GUIDE
STEP
Initializing
Test system and start the software
Setting / control parameters
ACTION
Install software and drivers
Connect sensor boards via USB
start the software
Set Reference current
Sensor calibration
Set IntegrationTime
Set Divider
Dark measurement
Calibration - Option
#1
Known target values
Calibration - Option
#2
Trial & Error
BASICS
Software and hardware are available, all system requirements are fulfilled
RESULT
Installed software
COMMENT /
DESCRIPTION
In/Decrease until the digits are in a range with an excellent ratio between noise (digits) and brightness digits
(several thousands, max. without saturation, in case of saturation consider also divider)
An ideal ratio between noise (digits) and brightness digits needs to be found (several thousands)
To prevent the risk of saturation increase divider as long as the sensor is not in saturation otherwise decrease gain or
IntegrationTime
Consider electronically noise – needed to increase accuracy
Use known XYZ values for the calibration target
Optimal digits = high accuracy
Optimal (max) digits
= high accuracy
Optimal (max) digits
= high accuracy
I
Dark
which correct sensor result ADC digit
Possible calibration for the existing target
Specify it as low as possible but so high as necessary
Specify it as low as possible but so high as necessary (see chapter
If possible decrease amplification to prevent saturation, consider the divider in the signal calculation
(see chapter 6.2)
see chapter 5.6 see chapter 5.6
Use a virtual calibration via simple calibration matrix
1 0 0
0 1 0
0 0 1
No existing reference values and spectrometer
No absolute measurement, estimation values for relative color measurement
ams.com db13386e_V2.01_certified_2017-07-06 Page 24 of 30
AS73210-AS89010-AB4 SET DK
V2.01
7 EXAMPLE FOR MEASUREMENT
The AS73210-AS89010-AB4 test system is ideal to detect and measure colors, color temperatures or brightness of lighting signals 8 . Typical applications are LED control and measurement and/or calibration of video walls and monitors.
Please note, the results was achieved by a standard and application specific calibrated test system (by using spectrometer and sensor, measured one by one, nearly at the same position). Measurements and calibration were done with a standard display under laboratory conditions. The following tables show the calculation of the calibration matrix and the results of the spectrometer, sensor and evaluation by using the set iref = 20nA, Tint = 1024ms and a RGB calibration.
Table 5: Example for calculation of the calibration matrix (calibration for RGB / target = 3)
Target data
T
Sensor data
S electronical offset
XYZ ADCOffset
S' = S - XYZ ADCOffset black-white offset
XYZ Offset
S'' = S' - XYZ Offset calculation: matrix for linear transformation K
X
Y
Z
X ADC
Y ADC
Z ADC
X ADCOffset
Y ADCOffset
Z ADCOffset
X' ADC
Y' ADC
Z' ADC
X offset
Y offset
Z offset
X'' ADC
Y'' ADC
Z'' ADC
A = T * S''_trans red green blue white grey 191 grey 127 grey 63 grey 31 yellow magenta cyan black
101,811295 89,8851089 44,6216393 241,16815 144,72527 73,803055 25,08864 7,8187065 192,6158 148,16277 136,15387 0,3010196
53,5799675 180,218567 19,6138611 259,89822 156,35645 80,397415 27,329342 8,4692755 235,91702 73,913986 202,53325 0,2844719
4,00652361 28,3652859 232,759201 272,6904 166,04439 85,696915 30,100254 9,0750713 32,101482 241,40384 264,61588 0,6896413
0,44342449 0,44284238 0,1933091 1,0995358 0,6618579 0,3389045 0,1158219 0,0363662 0,8936063 0,6407872 0,643008 0,0014194
0,23132649 0,81560093 0,10905106 1,1812601 0,7130977 0,367776 0,1255672 0,039099 1,0597086 0,3420032 0,9330936 0,0012544
0,02219289 0,06843875 0,43715331 0,5444769 0,3300425 0,1706812 0,0598643 0,0181949 0,0904617 0,4671361 0,5125724 0,0012142
0,002
0,00089
0,00122
0,44142449 0,44084238 0,1913091 1,0975358 0,6598579 0,3369045 0,1138219 0,0343662 0,8916063 0,6387872 0,641008 -0,000581
0,23043649 0,81471093 0,10816106 1,1803701 0,7122077 0,366886 0,1246772 0,038209 1,0588186 0,3411132 0,9322036 0,0003644
0,02097289 0,06721875 0,43593331 0,5432569 0,3288225 0,1694612 0,0586443 0,0169749 0,0892417 0,4659161 0,5113524 -5,83E-06
0,0
0,0
0,0
0,44
0,23
0,02
0,44
0,81
0,07
0,19
0,11
0,44
93,1036902 101,517741 27,6293008
106,851803 161,294271 21,7881287
58,8021521 49,2082394 103,458197
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
0,00
B = [S'' * S''_trans]^-1
10,8037949 -6,68471892 -3,1159837
-6,68471892 5,6283685 1,11105134
-3,11598367 1,11105134 6,4742213
K = A * B
241,163157 -20,295170 1,560052
8,306641 217,757058 -12,681089
-16,030909 -1,166386 541,257597
8 Note, the accuracy is depend on the system conditions, calibration and interferences. Therefore in application specific systems the sensor system must be optimized to increase accuracy. E.g. IR lighting will reduce the signal noise ratio dramatically. An add-on IR g will correct that. ams.com db13386e_V2.01_certified_2017-07-06 Page 25 of 30
AS73210-AS89010-AB4 SET DK
V2.01
Table 6: Calculations of sensor results
Spectrometer
X
Y
Z
X
Y
Z sensor ADC
Number of targets
1 red
2 green
3
blue
4 white
5 grey 191
6
grey 127
7 grey 63
8 grey 31
9
yellow
10 magenta
11 cyan
12 black
101,811295 89,8851089 44,6216393 241,168152 144,725266 73,8030548 25,0886402 7,81870651 192,615799 148,162766 136,15387 0,30101958
53,5799675 180,218567 19,6138611 259,898224 156,356445 80,3974152 27,3293419 8,46927547 235,917023 73,9139862 202,533249 0,2844719
4,00652361 28,3652859 232,759201 272,690399 166,044388 85,6969147 30,1002541 9,07507133 32,1014824 241,403839 264,615875 0,6896413 red green blue white grey 191 grey 127 grey 63 grey 31 yellow magenta cyan black
0,44342449 0,44284238 0,1933091 1,09953579 0,66185785 0,33890448 0,11582195 0,03636625 0,89360629 0,64078719 0,64300798 0,00141943
0,23132649 0,81560093 0,10905106 1,1812601 0,71309767 0,367776 0,1255672 0,03909899 1,0597086 0,34200324 0,93309362 0,00125439
0,02219289 0,06843875 0,43715331 0,5444769 0,33004246 0,17068117 0,05986427 0,01819494 0,09046171 0,46713612 0,51257241 0,00121417
Sensor XYZ after calibration
X
Y
Z
DIFF
Z
Spectrometerx
X
Y
Sensor x y y z z
DIFF delta Y (WP) delta xy
Spectrometeru' v'
Sensor u' v'
Diff u'v'
Number of targets
102,277
53,775
4,634
-0,466
-0,195
-0,627
0,639
0,336
0,025
0,637
0,335
0,029
0,075%
0,003
0,444
0,526
0,443
0,524
0,00122
3
90,351
180,414
28,993
-0,466
-0,195
-0,627
0,301
0,604
0,095
0,301
0,602
0,097
0,075%
0,002
0,125
0,564
0,125
0,563
0,00059
45,088
19,809
233,386
-0,466
-0,195
-0,627
0,150
0,066
0,784
0,151
0,066
0,782
0,075%
0,001
0,172
0,170
0,173
0,171
0,00123
242,043
259,457
275,698
-0,875
0,442
-3,007
0,312
0,336
0,352
0,311
0,334
0,355
0,170%
0,002
0,195
0,472
0,195
0,471
0,00125
145,658
156,595
167,196
-0,933
-0,238
-1,152
0,310
0,335
0,355
0,310
0,334
0,356
0,092%
0,001
0,194
0,471
0,194
0,470
0,00091
74,533
80,737
86,521
-0,730
-0,339
-0,824
0,308
0,335
0,357
0,308
0,334
0,358
0,130%
0,001
0,192
0,471
0,193
0,470
0,00102
25,477
27,546
30,399
-0,388
-0,217
-0,299
0,304
0,331
0,365
0,305
0,330
0,364
0,083%
0,002
0,191
0,468
0,192
0,468
0,00134
8,005
8,585
9,220
-0,186
-0,116
-0,144
0,308
0,334
0,358
0,310
0,333
0,357
0,045%
0,002
0,193
0,470
0,195
0,470
0,00181
194,139
237,035
33,402
-1,523
-1,118
-1,300
0,418
0,512
0,070
0,418
0,510
0,072
0,430%
0,002
0,201
0,555
0,202
0,554
0,00072
148,322
73,873
242,170
-0,159
0,041
-0,766
0,320
0,159
0,521
0,319
0,159
0,522
0,016%
0,000
0,299
0,336
0,299
0,335
0,00050
136,932
202,029
266,037
-0,778
0,504
-1,421
0,226
0,336
0,439
0,226
0,334
0,440
0,194%
0,002
0,137
0,459
0,138
0,459
0,00122
0,319
0,270
0,633
-0,018
0,015
0,057
0,236
0,223
0,541
0,261
0,221
0,518
0,006%
0,025
0,181
0,386
0,204
0,387
0,02230
See the in the lines “Delta”values the results as Delta Y, Delta xy or Delta u’v’ for the single colors, measured by the calibrated test system.
8 SOFTWARE LICENSE AGREEMENT
MATLAB ® . © 1984 - 2012 The MathWorks, Inc.
9 REFERENCE DOCUMENTS
[1] AS89010 Converter datasheet http://www.mazet.de/en/english-documents/english/data-sheets/as89010/download
[2] AS73210 Sensor datasheet http://www.mazet.de/en/english-documents/english/data-sheets/as73210/download
[3] EEPROM AT30TSE752 8751D–DTS–8/2012 Atmel datasheet http://www.atmel.com/Images/doc8751.pdf
[4] IOW24DG Converter I²C to USB - www.codemercs.com
[5] AS73210-AS89010-AB4 datasheet OEM Sensor board
[6] IO-Warrior Software http://www.codemercs.com/index.php?id=337&L=1 ams.com db13386e_V2.01_certified_2017-07-06 Page 26 of 30
AS73210-AS89010-AB4 SET DK
V2.01
10 MECHANICAL SIZES
Figure 25: Dimensions of the package and adapters
JCDK-Basic
JCDK-Aperture-13.5 with IR-cut-off filter
JCDK-1/4“-Adapter
JCDK-SMA-Adapter
ams.com db13386e_V2.01_certified_2017-07-06 Page 27 of 30
AS73210-AS89010-AB4 SET DK
V2.01
ORDERING INFORMATION
NAME STATUS
AS73210-AS89010-AB4 SET DK Development Kit
AS73210-AS89010-AB4 Series
Accessories JCDK-Aperture-13.5 with IR-cut-off filter
JCDK-1/4“-Adapter Accessories
ARTICLE
220800004
220800002 on request on request
JCDK-SMA-Adapter Accessories on request
The Development Kits and OEM modules described here are only to be used for compatibility-, qualification tests, and verification procedures or as demonstrators. They are not designed to be used in series products.
Please note, the OEM sensors and kits were designed to be a test system for usage in laboratories only.
They are not suitable for outdoor applications or to as series devices for inline measurements. Do not use the device in any process control or in any other series applications!
Before using engineering samples for anything other than system tests, please ask our sales team for status and availability of series or customized system solutions.
For more information please contact:
ams Sensors Germany GmbH:
Göschwitzer Straße 32
07745 JENA | GERMANY
Phone: +49 3641 2809-0
Fax: +49 3641 2809-12 [email protected] www.ams.com ams.com db13386e_V2.01_certified_2017-07-06 Page 28 of 30
AS73210-AS89010-AB4 SET DK
V2.01
LEGAL NOTES AND WARNINGS
Failure to comply with these instructions could result in death or serious injury.
Misuse of documentation – The information contained in this document is the property of ams
Sensors Germany. Photocopying or otherwise reproducing any part of the catalog, whether electronically or mechanically is prohibited, except where the express permission of ams Sensors
Germany GmbH has been obtained. In general, all company and brand names, as well as the names of individual products, are protected by brand, patent or product law.
State of document - The information provided in this document is for reference only. Do not use this document as product installation guide since products can be under development to improve performance or any other purpose. Before you start any development or place an order please contact your supplier or ams Sensors Germany for the latest version of this document. ams
Sensors Germany explicitly reserves the right to make technical changes to information described in the document.
Information and Disclaimer – The information provided in this document is based on the knowledge of the ams Sensors Germany GmbH as of the date of publication. The ams Sensors
Germany GmbH cannot give warranty regarding the accuracy of information provided by third parties. ams Sensors Germany may not have conducted testing or chemical analysis on all incoming material or chemicals. ams Sensors Germany GmbH performs and continues to perform reasonable measures to provide the most accurate data at the given time. Additional efforts to integrate information provided by third parties are performed and continue to be performed.
Certain supplier information may be proprietary or limited and not available at release.
Personal Injury: All products are conform to the specifications in accordance with the terms and conditions of ams Sensors Germanys standard warranty. Production processing does not necessarily include testing of all parameters.
RoHS Directive 2011/65/EU /REACH INFORMATION - RoHS compliance and PB free: The products of ams Sensors Germany fully comply with the current RoHS-directives. The ams
Sensors Germany semiconductor products do not contain any of the six substance chemical categories, for example including the restriction on lead usage (lead weight may not exceed
0.1% in homogeneous materials). RoHS compliant products are suitable for the usage in leadfree specified processes, when designed to be soldered at high temperatures.
REACH information: ams Sensors Germany products do not contain any of the latest REACH
Substances of Very High Concern (SVHC) regarding the Europe Union (EU) Regulation
1907/2006. The latest 155 substances restricted per the REACH Regulation were last updated on June 16, 2014. Please refer to the following for the most current candidate list of substances: http://echa.europa.eu/candidate-list-table. ams Sensors Germany solutions are not designed or intended for use in critical applications, in which the failures or malfunctions of the product may result in personal injury or death. Use of ams Sensors Germany products in life support systems is expressly unauthorized and any use by customer is completely at their own risk. In the case of a restricted use of the product described here, an application of the product outside of this limitation is at your own risk. ams.com db13386e_V2.01_certified_2017-07-06 Page 29 of 30
AS73210-AS89010-AB4 SET DK
V2.01
Warranty disclaimer – The warranty expressed herein shall be in lieu of any other warranties, expressed or implied, including, without limitation, any implied warranties or conditions of merchantability and fitness for a particular purpose., which are expressly disclaimed, and is in lieu of any and all other obligations or liability on supplier’s part- For the avoidance of doubt, supplier shall not be liable for any special, incidental, indirect or consequential loss or damage, including loss pf revenue or profit, of any kind of nature, arising at any time, from any cause whatsoever resulting from the use or operation of the products or any breach of this limited warranty.
Legal liability - ams Sensors Germany assumes no responsibility for the use of any foreign products or circuits described in this document or customer product design, conveys no license, either expressed or implied, under any patent or other right, and makes no representation that the foreign circuits are free of patent infringement. ams Sensors Germany further makes no claim as to the suitability of its products for any particular purpose, nor does ams Sensors
Germany assume any liability arising out of the use of any product or circuit, and specifically disclaims any and all liability, including without limitation consequential or incidental damages.
ESD Warning: Sensor handling precautions should be observed to avoid static discharge.
WEEE Disposal: - The product should be disposed in to according the Directive 2002/96 / EC of the European Council on Waste Electrical and Electronic Equipment [WEEE] and the German electoral law [ElektroG] of 16 March 2005. Please contact our technical support if you need more details. ams.com db13386e_V2.01_certified_2017-07-06 Page 30 of 30