Mindray Passport 12m and 17m Operator’s Manual

Add to My manuals
410 Pages

advertisement

Mindray Passport 12m and 17m Operator’s Manual | Manualzz

Passport 12m/Passport 17m

Patient Monitor

Operator’s Manual

© Copyright 2013 - 2019 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved.

Release time: 2019-0 8

Revision: 1 7 .0

WARNING

Federal Law (USA) restricts this device to sale by or on the order of a physician or other practitioner licensed by U.S. state law to use or order the use of this device.

Intellectual Property Statement

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this Mindray product and this manual. This manual may refer to information protected by copyrights or patents and does not convey any license under the patent rights or copyright of Mindray, or of others.

Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.

Release, amendment, reproduction, distribution, rental, adaption and translation or any other derivative work of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.

, , and are the trademarks, registered or otherwise, of Mindray in China and other countries. All other trademarks that appear in this manual are used only for information or editorial purposes. They are the property of their respective owners.

This posting serves as notice under 35 U.S.C.

§

287(a) for Mindray patents: http://www.mindrayna.com/patents.

Passport 12m/Passport 17m Operator's Manual I

Manufacturer’s Responsibility

Contents of this manual are subject to changes without prior notice.

All information contained in this manual is believed to be correct. Mindray is not liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing, performance, or use of this manual.

Mindray is responsible for the effects on safety, reliability and performance of this product, only if:

 all installation operations, expansions, changes, modifications and repairs of this product are conducted by Mindray authorized personnel;

 the electrical installation of the relevant room complies with the applicable national and local requirements;

 the product is used in accordance with the instructions for use.

WARNING

Only skilled/trained clinical professionals should operate this equipment.

It is important for the hospital or organization that uses this equipment to perform a recommended service/maintenance plan. Neglect of this may result in machine breakdown or personal injury.

II Passport 12m/Passport 17m Operator's Manual

Service

Mindray maintains a network of service representatives and factory-trained distributors. Prior to requesting service, perform a complete operational check of the instrument to verify proper control settings. If operational problems continue to exist, contact Mindray Service.

In North America contact the Service Department at (800) 288-2121, ext: 8116 for Technical Support or (201) 995-8000 for assistance in determining the nearest field service location.

Please include the instrument model number, the serial number, and a description of the problem with all requests for service.

Any questions regarding the warranty should be directed to your local sales or service representative.

Company Contact

Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

Address:

Website:

E-mail Address:

Tel:

Fax:

Distributor:

Address:

Mindray Building, Keji 12th Road South, High-tech industrial park, Nanshan,Shenzhen

518057,P.R.China www.mindray.com [email protected]

+86 755 81888998

+86 755 26582680

Mindray DS USA, Inc.

800 MacArthur Boulevard, Mahwah, New Jersey, 07430 USA

Tel:

Website:

1.800.288.2121, 1.201.995.8000 www.mindray.com

Passport 12m/Passport 17m Operator's Manual III

Preface

Manual Purpose

This manual contains the instructions necessary to operate the product safely and in accordance with its function and intended use. Observance of this manual is a prerequisite for proper product performance and correct operation and ensures patient and operator safety.

This manual is based on the maximum configuration and therefore some contents may not apply to your product. If you have any question, please contact Mindray.

This manual is an integral part of the product. It should always be kept close to the equipment so that it can be conveniently referenced when needed.

Intended Audience

This manual is intended for clinical professionals who are expected to have a working knowledge of medical procedures, practices and terminology as required for monitoring of patients.

Illustrations

All illustrations in this manual serve as examples only. They may not necessarily reflect the setup or data displayed on your monitor.

Conventions

Italic text

is used to quote the referenced chapters or sections.

[ ] is used to enclose screen texts.

→ is used to indicate operational procedures.

IV Passport 12m/Passport 17m Operator's Manual

Contents

1 Safety.............................................................................................................................................................................. 1-1

1.1 Safety Information .................................................................................................................................................................................... 1-1

1.1.1 Warnings ........................................................................................................................................................................................ 1-1

1.1.2 Cautions ......................................................................................................................................................................................... 1-2

1.1.3 Notes ............................................................................................................................................................................................... 1-3

1.2 Equipment Symbols ................................................................................................................................................................................. 1-3

2 The Basics ....................................................................................................................................................................... 2-1

2.1 Monitor Description ................................................................................................................................................................................. 2-1

2.1.1 Intended Use ................................................................................................................................................................................ 2-1

2.1.2 Applied Parts ................................................................................................................................................................................ 2-2

2.2 Main Unit ..................................................................................................................................................................................................... 2-2

2.2.1 Front View ...................................................................................................................................................................................... 2-2

2.2.2 Side View ........................................................................................................................................................................................ 2-4

2.2.3 Rear View ....................................................................................................................................................................................... 2-5

2.2.4 Bottom View (Passport 17m) ................................................................................................................................................... 2-7

2.3 Satellite Module Rack .............................................................................................................................................................................. 2-8

2.4 Modules ..................................................................................................................................................................................................... 2-10

2.4.1 Multi-Parameter Module ......................................................................................................................................................... 2-11

2.4.2 T1 .................................................................................................................................................................................................... 2-13

2.4.3 N1 ................................................................................................................................................................................................... 2-14

2.5 Display Screen ......................................................................................................................................................................................... 2-1 5

2.6 QuickKeys .................................................................................................................................................................................................. 2-1 7

3 Basic Operations ............................................................................................................................................................ 3-1

3.1 Installation .................................................................................................................................................................................................. 3-1

3.1.1 Unpacking and Checking ......................................................................................................................................................... 3-1

3.1.2 Environmental Requirements ................................................................................................................................................. 3-2

3.2 Getting Started .......................................................................................................................................................................................... 3-2

3.2.1 Connecting to Power Source ................................................................................................................................................... 3-2

3.2.2 Turning Power On ....................................................................................................................................................................... 3-3

3.2.3 Starting Monitoring .................................................................................................................................................................... 3-4

3.3 Turning Off the Monitor .......................................................................................................................................................................... 3-4

3.4 Using the Knob .......................................................................................................................................................................................... 3-5

3.5 Using a Mouse ........................................................................................................................................................................................... 3-5

3.6 Using Keys ................................................................................................................................................................................................... 3-5

3.7 Using the On-screen Keyboard ............................................................................................................................................................ 3-6

3.8 Using the Touchscreen ............................................................................................................................................................................ 3-6

3.9 Using the [Main Menu] ........................................................................................................................................................................... 3-6

3.10 Setting the Screen .................................................................................................................................................................................. 3-7

3.11 Displaying the Timer .............................................................................................................................................................................. 3-8

3.12 Changing General Settings ................................................................................................................................................................. 3-9

3.12.1 Setting up a Monitor ................................................................................................................................................................ 3-9

Passport 12m/Passport 17m Operator's Manual 1

3.12.2 Changing Language ................................................................................................................................................................ 3-9

3.12.3 Adjusting the Screen Brightness ........................................................................................................................................ 3-10

3.12.4 Showing/Hiding the Help Text ........................................................................................................................................... 3-10

3.12.5 Setting the Date and Time ................................................................................................................................................... 3-10

3.12.6 Adjusting Volume ................................................................................................................................................................... 3-10

3.13 Setting Parameters .............................................................................................................................................................................. 3-11

3.13.1 Switching the Parameters On/Off ...................................................................................................................................... 3-11

3.13.2 Accessing the Parameters Menu ........................................................................................................................................ 3-12

3.13.3 Removing a Module Conflict ............................................................................................................................................... 3-12

3.14 Using a CF Storage Card ..................................................................................................................................................................... 3-13

3.15 Operating Modes ................................................................................................................................................................................. 3-14

3.15.1 Monitoring Mode.................................................................................................................................................................... 3-14

3.15.2 Night Mode ............................................................................................................................................................................... 3-14

3.15.3 Privacy Mode ............................................................................................................................................................................ 3-15

3.15.4 Standby Mode ......................................................................................................................................................................... 3-15

4 Managing Patients ......................................................................................................................................................... 4-1

4.1 Admitting a Patient .................................................................................................................................................................................. 4-1

4.2 Quick Admitting a Patient ...................................................................................................................................................................... 4-2

4.3 Setting the Monitor Location ................................................................................................................................................................ 4-2

4.4 Querying and Obtaining Patient Information ................................................................................................................................. 4-2

4.5 Querying from Local Facility ................................................................................................................................................................. 4-3

4.6 Associating Patient Information ........................................................................................................................................................... 4-3

4.7 Editing Patient Information ................................................................................................................................................................... 4-3

4.8 Discharging a Patient .............................................................................................................................................................................. 4-4

4.9 Transferring Patient Data ........................................................................................................................................................................ 4-4

4.9.1 Transferring Patient Data via MPM/T1/N1 ........................................................................................................................... 4-5

4.9.2 Transferring Patient Data via a CF Storage Card or USB Drive ....................................................................................... 4-6

4.10 Connecting to a CMS............................................................................................................................................................................. 4-7

5 Managing Configurations ............................................................................................................................................. 5-1

5.1 Introduction ............................................................................................................................................................................................... 5-1

5.2 Accessing the [Manage Configuration] Menu ................................................................................................................................. 5-2

5.3 Changing Department ............................................................................................................................................................................ 5-2

5.4 Setting Default Configuration .............................................................................................................................................................. 5-3

5.5 Saving Current Settings .......................................................................................................................................................................... 5-3

5.6 Editing Configurations ............................................................................................................................................................................ 5-4

5.7 Deleting a Configuration ........................................................................................................................................................................ 5-5

5.8 Transferring a Configuration ................................................................................................................................................................. 5-5

5.9 Loading a Configuration ......................................................................................................................................................................... 5-5

5.10 Restoring the Latest Configuration Automatically ....................................................................................................................... 5-6

6 User Screens ................................................................................................................................................................... 6-1

6.1 Configuring Your Screens ....................................................................................................................................................................... 6-1

6.1.1 Changing the Waveform Line Size ......................................................................................................................................... 6-1

6.1.2 Changing Measurement Colors .............................................................................................................................................. 6-1

2 Passport 12m/Passport 17m Operator's Manual

6.1.3 Changing Screen Layout ........................................................................................................................................................... 6-1

6.1.4 Setting the Waveform Sweep Mode ...................................................................................................................................... 6-2

6.2 Viewing Minitrends .................................................................................................................................................................................. 6-2

6.2.1 Having a Split-Screen View of Minitrends............................................................................................................................ 6-2

6.2.2 Setting Minitrends ...................................................................................................................................................................... 6-3

6.3 Viewing OxyCRG ....................................................................................................................................................................................... 6-3

6.4 Viewing Other Patients ........................................................................................................................................................................... 6-4

6.4.1 Care Group .................................................................................................................................................................................... 6-4

6.4.2 Viewing the Care Group Overview Bar ................................................................................................................................. 6-5

6.4.3 Understanding the View Other Patient Window ............................................................................................................... 6-6

6.5 Understanding the Big Numerics Screen .......................................................................................................................................... 6-7

7 Alarms............................................................................................................................................................................. 7-1

7.1 Alarm Categories ....................................................................................................................................................................................... 7-1

7.2 Alarm Levels ............................................................................................................................................................................................... 7-2

7.3 Alarm Indicators ........................................................................................................................................................................................ 7-2

7.3.1 Alarm Lamp ................................................................................................................................................................................... 7-2

7.3.2 Alarm Messages ........................................................................................................................................................................... 7-2

7.3.3 Flashing Numeric ........................................................................................................................................................................ 7-3

7.3.4 Audible Alarm Tones ................................................................................................................................................................... 7-3

7.3.5 Alarm Status Symbols ................................................................................................................................................................ 7-4

7.4 Alarm Tone Configuration ...................................................................................................................................................................... 7-4

7.4.1 Setting the Minimum Alarm Volume..................................................................................................................................... 7-4

7.4.2 Changing the Alarm Volume ................................................................................................................................................... 7-4

7.4.3 Setting the Interval Between Alarm Sounds ....................................................................................................................... 7-5

7.4.4 Changing the Alarm Tone Pattern .......................................................................................................................................... 7-5

7.4.5 Setting the Reminder Tones..................................................................................................................................................... 7-6

7.5 Understanding the Alarm Setup Menu .............................................................................................................................................. 7-6

7.5.1 Setting Alarm Properties for All Parameters ........................................................................................................................ 7-7

7.5.2 Adjusting Alarm Limits Automatically ................................................................................................................................... 7-7

7.5.3 Setting Alarm Delay Time ....................................................................................................................................................... 7-10

7.5.4 Setting SpO2 Technical Alarm Delay ................................................................................................................................... 7-11

7.5.5 Setting Recording Length ...................................................................................................................................................... 7-11

7.5.6 Entering CPB Mode (Cardiopulmonary Bypass Mode) .................................................................................................. 7-11

7.6 Pausing Alarms ........................................................................................................................................................................................ 7-12

7.7 Switching Off All Alarms ....................................................................................................................................................................... 7-12

7.8 Resetting Alarms ..................................................................................................................................................................................... 7-13

7.9 Latching Alarms ....................................................................................................................................................................................... 7-14

7.10 Testing Alarms ....................................................................................................................................................................................... 7-15

7.11 Using Care Group Alarms ................................................................................................................................................................... 7-15

7.11.1 Care Group Auto Alarms ........................................................................................................................................................ 7-15

7.11.2 Setting Care Group Alert Tone............................................................................................................................................. 7-15

7.11.3 Resetting Care Group Alarms .............................................................................................................................................. 7-16

8 Monitoring ECG .............................................................................................................................................................. 8-1

8.1 Introduction ............................................................................................................................................................................................... 8-1

Passport 12m/Passport 17m Operator's Manual 3

4

8.2 Safety ............................................................................................................................................................................................................ 8-1

8.3 Preparing to Monitor ECG ...................................................................................................................................................................... 8-2

8.3.1 Preparing the Patient and Placing the Electrodes ............................................................................................................ 8-2

8.3.2 Choosing AHA or IEC Lead Placement .................................................................................................................................. 8-2

8.3.3 ECG Lead Placements ................................................................................................................................................................ 8-3

8.3.4 Checking Paced Status .............................................................................................................................................................. 8-4

8.4 Understanding the ECG Display ........................................................................................................................................................... 8-5

8.5 Changing ECG Settings ........................................................................................................................................................................... 8-6

8.5.1 Accessing ECG Menus ................................................................................................................................................................ 8-6

8.5.2 Choosing the Alarm Source ..................................................................................................................................................... 8-6

8.5.3 Changing ECG Wave Settings .................................................................................................................................................. 8-6

8.5.4 Changing the ECG Filter Settings ........................................................................................................................................... 8-7

8.5.5 Setting the ECG Lead Set .......................................................................................................................................................... 8-7

8.5.6 Choosing an ECG Display Screen ........................................................................................................................................... 8-7

8.5.7 Setting the Notch Filter ............................................................................................................................................................. 8-8

8.5.8 Changing the Pacer Reject Settings ...................................................................................................................................... 8-8

8.5.9 Enabling Smart Lead Off ........................................................................................................................................................... 8-8

8.5.10 Setting the Alarm Level for ECG Lead Off Alarms ............................................................................................................ 8-8

8.5.11 Adjusting QRS Volume ............................................................................................................................................................. 8-9

8.5.12 About the Defibrillator Synchronization ........................................................................................................................... 8-9

8.5.13 Adjusting the Minimum QRS Detection Threshold (For Advanced ECG Algorithm Only) .................................. 8-9

8.6 About ST Monitoring ............................................................................................................................................................................. 8-10

8.6.1 Switching ST Monitoring On and Off .................................................................................................................................. 8-10

8.6.2 Understanding the ST Display ............................................................................................................................................... 8-11

8.6.3 Saving the Current ST Segment as Reference .................................................................................................................. 8-12

8.6.4 Changing the Reference Segment ....................................................................................................................................... 8-12

8.6.5 Deleting a Reference Segment ............................................................................................................................................. 8-12

8.6.6 Recording the ST Segment ..................................................................................................................................................... 8-12

8.6.7 Changing the ST Alarm Limits ............................................................................................................................................... 8-13

8.6.8 Setting the ST Alarm Delay Time .......................................................................................................................................... 8-13

8.6.9 Adjusting ST Measurement Points ....................................................................................................................................... 8-13

8.7 QT/QTc Interval Monitoring (For Advanced ECG Algorithm Only) ........................................................................................... 8-14

8.7.1 QT/QTc Monitoring Limitations ............................................................................................................................................ 8-15

8.7.2 Enabling QT/QTc Monitoring ................................................................................................................................................. 8-15

8.7.3 Displaying QT/QTc Parameters and Waveform ................................................................................................................ 8-16

8.7.4 Accessing the QT View ............................................................................................................................................................. 8-16

8.7.5 Changing QT Settings .............................................................................................................................................................. 8-17

8.8 About Arrhythmia Monitoring ............................................................................................................................................................ 8-18

8.8.1 Understanding the Arrhythmia Events ............................................................................................................................... 8-19

8.8.2 Changing Arrhythmia Alarm Settings ................................................................................................................................. 8-20

8.8.3 Changing Arrhythmia Threshold Settings ......................................................................................................................... 8-20

8.8.4 Setting the Extended Arrhythmia (For Advanced ECG Algorithm Only) .................................................................. 8-21

8.8.5 Reviewing Arrhythmia Events ............................................................................................................................................... 8-21

8.9 ECG Relearning ........................................................................................................................................................................................ 8-21

8.9.1 Initiating an ECG Relearning Manually ............................................................................................................................... 8-21

8.9.2 Automatic ECG Relearning ..................................................................................................................................................... 8-22

Passport 12m/Passport 17m Operator's Manual

8.10 12-lead ECG Monitoring ..................................................................................................................................................................... 8-22

8.10.1 Setting ECG Waveform Sequence ...................................................................................................................................... 8-23

8.10.2 Extending the Rhythm Lead Waveform area .................................................................................................................. 8-23

8.11 Resting 12-Lead ECG Analysis ........................................................................................................................................................... 8-23

8.11.1 Accessing the 12-Lead Screen ............................................................................................................................................. 8-23

8.11.2 Entering Patient Information ............................................................................................................................................... 8-24

8.11.3 12-Lead Setup .......................................................................................................................................................................... 8-25

8.11.4 Setting the 12-lead Order.......

............................................................................................................................................... 8-26

8.11.

5 Resting 12-lead ECG Analysis............................................................................................................................................... 8-26

8.11.

6 12-lead ECG Report ................................................................................................................................................................ 8-28

8.12 Troubleshooting ................................................................................................................................................................................... 8-29

9 Monitoring Respiration (Resp) ...................................................................................................................................... 9-1

9.1 Introduction ............................................................................................................................................................................................... 9-1

9.2 Safety Information .................................................................................................................................................................................... 9-1

9.3 Understanding the Resp Display ......................................................................................................................................................... 9-1

9.4 Placing Resp Electrodes .......................................................................................................................................................................... 9-2

9.4.1 Optimizing Lead Placement for Resp .................................................................................................................................... 9-3

9.4.2 Cardiac Overlay ............................................................................................................................................................................ 9-3

9.4.3 Abdominal Breathing ................................................................................................................................................................. 9-3

9.4.4 Lateral Chest Expansion ............................................................................................................................................................ 9-3

9.5 Choosing the Respiration Lead ............................................................................................................................................................ 9-3

9.6 Changing the Apnea Alarm Delay ....................................................................................................................................................... 9-3

9.7 Changing Resp Detection Mode .......................................................................................................................................................... 9-4

9.8 Changing Resp Wave Settings .............................................................................................................................................................. 9-4

9.9 Setting Respiration Rate (RR) Source .................................................................................................................................................. 9-5

9.10 Setting Alarm Properties ...................................................................................................................................................................... 9-5

10 Monitoring PR ........................................................................................................................................................... 10-1

10.1 Introduction ........................................................................................................................................................................................... 10-1

10.2 Setting the PR Source ......................................................................................................................................................................... 10-1

10.3 Selecting the Active Alarm Source .................................................................................................................................................. 10-2

10.4 QRS Tone ................................................................................................................................................................................................. 10-2

11 Monitoring SpO

2

........................................................................................................................................................ 11-1

11.1 Introduction ........................................................................................................................................................................................... 11-1

11.2 Safety ........................................................................................................................................................................................................ 11-2

11.3 Identifying SpO

2

Modules .................................................................................................................................................................. 11-2

11.4 Applying the Sensor ............................................................................................................................................................................ 11-2

11.5 Changing SpO

2

Settings ..................................................................................................................................................................... 11-3

11.5.1 Accessing SpO2 Menus ......................................................................................................................................................... 11-3

11.5.2 Adjusting the Desat Alarm .................................................................................................................................................... 11-3

11.5.3 Setting SpO2 Sensitivity........................................................................................................................................................ 11-3

11.5.4 Changing Averaging Time .................................................................................................................................................... 11-3

11.5.5 Monitoring SpO2 and NIBP Simultaneously ................................................................................................................... 11-4

11.5.6 Sat-Seconds Alarm Management ....................................................................................................................................... 11-4

11.5.7 Changing the Speed of the Pleth/Plethb Wave ............................................................................................................. 11-5

Passport 12m/Passport 17m Operator's Manual 5

11.5.8 Zooming PI Value .................................................................................................................................................................... 11-5

11.5.9 Setting the Alarm Level for SpO2 Sensor Off Alarm ..................................................................................................... 11-5

11.5.10 Setting the SpO2 Tone Mode ............................................................................................................................................ 11-6

11.6 Measurement Limitations .................................................................................................................................................................. 11-6

11.7 Masimo Information ............................................................................................................................................................................ 11-7

11.8 Nellcor Information .............................................................................................................................................................................. 11-7

11.9 Troubleshooting .................................................................................................................................................................................... 11-8

12 Monitoring NIBP ....................................................................................................................................................... 12-1

12.1 Introduction ........................................................................................................................................................................................... 12-1

12.2 Safety ....................................................................................................................................................................................................... 12-2

12.3 Measurement Limitations .................................................................................................................................................................. 12-2

12.4 Measurement Methods ...................................................................................................................................................................... 12-3

12.5 Setting Up the NIBP Measurement ................................................................................................................................................. 12-3

12.5.1 Preparing the Patient ............................................................................................................................................................. 12-3

12.5.2 Preparing to Measure NIBP .................................................................................................................................................. 12-3

12.5.3 Starting and Stopping Measurements ............................................................................................................................. 12-4

12.5.4 Correcting the Measurement if Limb is not at Heart Level ........................................................................................ 12-4

12.5.5 Enabling NIBP Auto Cycling and Setting the Interval .................................................................................................. 12-4

12.5.6 Enabling Measurement on Clock ....................................................................................................................................... 12-4

12.5.7 Starting a STAT Measurement ............................................................................................................................................. 12-5

12.6 Understanding the NIBP Numerics ................................................................................................................................................. 12-5

12.7 Changing NIBP Settings ..................................................................................................................................................................... 12-6

12.7.1 Setting the Initial Cuff Inflation Pressure ......................................................................................................................... 12-6

12.7.2 Setting NIBP Alarm Properties ............................................................................................................................................ 12-6

12.7.3 Displaying NIBP List ............................................................................................................................................................... 12-6

12.7.4 Setting the Pressure Unit ...................................................................................................................................................... 12-6

12.7.5 Switching On NIBP End Tone ............................................................................................................................................... 12-6

12.8 Assisting Venous Puncture ................................................................................................................................................................ 12-7

13 Monitoring Temp ...................................................................................................................................................... 13-1

13.1 Introduction ........................................................................................................................................................................................... 13-1

13.2 Safety ....................................................................................................................................................................................................... 13-1

13.3 Taking a Temp Measurement ............................................................................................................................................................ 13-1

13.4 Understanding the Temp Display .................................................................................................................................................... 13-1

13.5 Setting the Temperature Unit ........................................................................................................................................................... 13-2

14 Monitoring IBP .......................................................................................................................................................... 14-1

14.1 Introduction ........................................................................................................................................................................................... 14-1

14.2 Safety ....................................................................................................................................................................................................... 14-1

14.3 Zeroing the Transducer ...................................................................................................................................................................... 14-2

14.4 Setting Up the Pressure Measurement .......................................................................................................................................... 14-3

14.5 Understanding the IBP Display ........................................................................................................................................................ 14-4

14.6 Changing IBP Settings ........................................................................................................................................................................ 14-4

14.6.1 Changing a Pressure for Monitoring ................................................................................................................................. 14-4

14.6.2 Setting Alarm Properties ...................................................................................................................................................... 14-5

6 Passport 12m/Passport 17m Operator's Manual

14.6.3 Changing Averaging Time .................................................................................................................................................... 14-5

14.6.4 Setting Up the IBP Wave ....................................................................................................................................................... 14-5

14.6.5 Setting the Pressure Unit ...................................................................................................................................................... 14-5

14.6.6 Enabling PPV Measurement and Setting PPV Source ................................................................................................. 14-6

14.7 Overlapping IBP Waveforms ............................................................................................................................................................. 14-6

14.8 Measuring PAWP ................................................................................................................................................................................... 14-7

14.8.1 Preparing to Measure PAWP ................................................................................................................................................ 14-7

14.8.2 Setting Up the PAWP Measurement ................................................................................................................................. 14-8

14.8.3 Understanding the PAWP Setup Menu ............................................................................................................................ 14-9

14.9 Troubleshooting ................................................................................................................................................................................ 14-10

15 Monitoring Cardiac Output ...................................................................................................................................... 15-1

15.1 Introduction ........................................................................................................................................................................................... 15-1

15.2 Understanding the C.O. Display ....................................................................................................................................................... 15-2

15.3 Influencing Factors .............................................................................................................................................................................. 15-2

15.4 Setting Up the C.O. Measurement .................................................................................................................................................. 15-3

15.5 Measuring the Blood Temperature ................................................................................................................................................. 15-6

15.6 Changing C.O. Settings ....................................................................................................................................................................... 15-6

15.6.1 Setting the Temperature Unit ............................................................................................................................................. 15-6

15.6.2 Setting Alarm Properties ...................................................................................................................................................... 15-6

16 Monitoring CCO/SvO

2

............................................................................................................................................... 16-1

16.1 Introduction ........................................................................................................................................................................................... 16-1

16.2 Safety ....................................................................................................................................................................................................... 16-1

16.3 Automatic Communication Detection .......................................................................................................................................... 16-2

16.4 Connecting the Device ....................................................................................................................................................................... 16-2

16.4.1 Connecting the Vigilance II® Monitor ............................................................................................................................... 16-2

16.4.2 Connecting the Vigileo

TM

Monitor ..................................................................................................................................... 16-4

16.4.3 Connecting the EV1000 Monitor ....................................................................................................................................... 16-5

16.5 Understanding CCO Parameters ...................................................................................................................................................... 16-6

16.5.1 Hemodynamic Parameters for Vigilance II® Monitor ................................................................................................... 16-6

16.5.2 Hemodynamic Parameters for Vigileo TM Monitor ......................................................................................................... 16-7

16.5.3 Hemodynamic Parameters for EV1000 Monitor ............................................................................................................ 16-7

16.6 Understanding the CCO Display ...................................................................................................................................................... 16-8

16.7 Changing CCO Settings ...................................................................................................................................................................... 16-9

16.7.1 Selecting Vascular Resistance Unit .................................................................................................................................... 16-9

16.7.2 Selecting the Displayed Parameters ................................................................................................................................. 16-9

16.7.3 Checking the C.O. Measurements ..................................................................................................................................... 16-9

16.7.4 Setting Signal Output ......................................................................................................................................................... 16-10

16.7.5 Selecting Alarm Properties ............................................................................................................................................... 16-10

16.8 Understanding SvO

2

/ScvO

2

Parameters ...................................................................................................................................... 16-11

16.8.1 Oxygenation Parameters for Vigilance II® Monitor ..................................................................................................... 16-11

16.8.2 Oxygenation Parameters for Vigileo

TM

Monitor ........................................................................................................... 16-11

16.8.3 Oxygenation Parameters for EV1000 Monitor ............................................................................................................ 16-12

16.9 Understanding the SvO

2

/ScvO

2

Display ..................................................................................................................................... 16-12

16.10 Changing SvO

2

/ScvO

2

Settings ................................................................................................................................................... 16-13

Passport 12m/Passport 17m Operator's Manual 7

16.10.1 Setting Signal Output ...................................................................................................................................................... 16-13

16.10.2 Selecting Alarm Properties ............................................................................................................................................. 16-13

17 Monitoring ScvO

2

...................................................................................................................................................... 17-1

17.1 Introduction ........................................................................................................................................................................................... 17-1

17.2 Safety Information ............................................................................................................................................................................... 17-1

17.3 Performing ScvO

2

Measurements ................................................................................................................................................... 17-2

17.4 ScvO

2

Calibration .................................................................................................................................................................................. 17-3

17.5 Understanding the ScvO

2

Display ................................................................................................................................................... 17-3

17.6 Understanding ScvO

2

Parameters ................................................................................................................................................... 17-4

17.7 Changing ScvO

2

Settings ................................................................................................................................................................... 17-4

17.7.1 Selecting Hb/Hct ..................................................................................................................................................................... 17-4

17.7.2 Selecting Alarm Properties .................................................................................................................................................. 17-4

18 Monitoring CO

2

......................................................................................................................................................... 18-1

18.1 Introduction ........................................................................................................................................................................................... 18-1

18.2 Identifying CO

2

Modules .................................................................................................................................................................... 18-2

18.3 Preparing to Measure CO

2

................................................................................................................................................................. 18-2

18.3.1 Making a Sidestream CO2 Measurement ........................................................................................................................ 18-3

18.3.2 Making a Microstream CO

2

Measurement ...................................................................................................................... 18-4

18.4 Changing CO

2

Settings ....................................................................................................................................................................... 18-4

18.4.1 Accessing CO2 Menus ........................................................................................................................................................... 18-4

18.4.2 Setting the CO

2

Unit ............................................................................................................................................................... 18-4

18.4.3 Setting Up Gas Compensations ......................................................................................................................................... 18-5

18.4.4 Setting Up Humidity Compensation ................................................................................................................................ 18-5

18.4.5 Setting the Apnea Alarm Delay .......................................................................................................................................... 18-6

18.4.6 Choosing a Time Interval for Peak-Picking...................................................................................................................... 18-6

18.4.7 Setting the Flow Rate ............................................................................................................................................................ 18-6

18.4.8 Setting up the CO2 Wave ..................................................................................................................................................... 18-7

18.4.9 Setting RR Source ................................................................................................................................................................... 18-7

18.4.10 Barometric Pressure Compensation ............................................................................................................................... 18-7

18.4.11 Entering the Standby Mode .............................................................................................................................................. 18-7

18.5 Measurement Limitations .................................................................................................................................................................. 18-8

18.6 Leakage test ........................................................................................................................................................................................... 18-8

18.7 Troubleshooting the Sidestream CO

2

Sampling System .......................................................................................................... 18-9

18.8 Removing Exhaust Gases from the System .................................................................................................................................. 18-9

18.9 Zeroing the Sensor .............................................................................................................................................................................. 18-9

18.10 Calibrating the Sensor ................................................................................................................................................................... 18-10

18.11 Oridion Information ....................................................................................................................................................................... 18-10

19 Monitoring AG ........................................................................................................................................................... 19-1

19.1 Introduction ........................................................................................................................................................................................... 19-1

19.2 Identifying AG Modules ...................................................................................................................................................................... 19-1

19.3 Understanding the AG Display ......................................................................................................................................................... 19-2

19.4 MAC Values ............................................................................................................................................................................................. 19-3

19.5 Preparing to Measure AG ................................................................................................................................................................... 19-4

8 Passport 12m/Passport 17m Operator's Manual

19.6 Changing AG Settings ......................................................................................................................................................................... 19-5

19.6.1 Setting Gas Unit ...................................................................................................................................................................... 19-5

19.6.2 Setting the Apnea Alarm Delay .......................................................................................................................................... 19-5

19.6.3 Changing the Sample Flow Rate ........................................................................................................................................ 19-6

19.6.4 Entering the Standby Mode ................................................................................................................................................ 19-6

19.6.5 Setting Up the AG Wave ........................................................................................................................................................ 19-7

19.6.6 Setting RR Source ................................................................................................................................................................... 19-7

19.7 Changing the Anesthetic Agent ....................................................................................................................................................... 19-7

19.8 Measurement Limitations .................................................................................................................................................................. 19-7

19.9 Troubleshooting ................................................................................................................................................................................... 19-8

19.9.1 When the Gas Inlet is Blocked ............................................................................................................................................. 19-8

19.9.2 When an Internal Occlusion Occurs .................................................................................................................................. 19-8

19.10 Removing Exhaust Gases from the System................................................................................................................................ 19-8

20 Monitoring Bispectral Index (BIS) ............................................................................................................................ 20-1

20.1 Introduction ........................................................................................................................................................................................... 20-1

20.2 Safety Information ............................................................................................................................................................................... 20-2

20.3 Understanding the BIS Display ........................................................................................................................................................ 20-3

20.3.1 BIS Parameter Area ................................................................................................................................................................. 20-3

20.3.2 BIS Waveform Area.................................................................................................................................................................. 20-5

20.3.3 BIS Expand View ...................................................................................................................................................................... 20-6

20.4 Setting up the BIS Measurement ..................................................................................................................................................... 20-8

20.5 Auto Impedance Check ...................................................................................................................................................................... 20-9

20.6 Sensor Check ......................................................................................................................................................................................... 20-9

20.7 BIS Sensor Check Window .............................................................................................................................................................. 20-10

20.8 Choosing the BIS Smoothing Rate ................................................................................................................................................ 20-11

20.9 Changing the Secondary Parameters .......................................................................................................................................... 20-11

20.10 Changing the EEG Wave Size ........................................................................................................................................................ 20-11

20.11 Changing the Speed of the EEG Wave ....................................................................................................................................... 20-11

20.12 Setting the Trend Length .............................................................................................................................................................. 20-12

20.13 Switching the Filter On or Off ...................................................................................................................................................... 20-12

20.14 Covidien Information ..................................................................................................................................................................... 20-12

21 Monitoring NMT (From TOF-Watch® SX Monitor) .................................................................................................... 21-1

21.1 Introduction ........................................................................................................................................................................................... 21-1

21.2 Safety ....................................................................................................................................................................................................... 21-1

21.3 Connecting a TOF-Watch® SX monitor .......................................................................................................................................... 21-1

21.4 NMT Parameters.................................................................................................................................................................................... 21-1

21.5 Accessing the NMT Setup menu ...................................................................................................................................................... 21-2

21.6 NMT Display ........................................................................................................................................................................................... 21-2

22 Monitoring Respiratory Mechanics (RM) ................................................................................................................. 22-1

22.1 Introduction ........................................................................................................................................................................................... 22-1

22.2 Safety Information ............................................................................................................................................................................... 22-2

22.3 Preparing to Monitor RM ................................................................................................................................................................... 22-3

22.4 Understanding the RM Display ........................................................................................................................................................ 22-4

Passport 12m/Passport 17m Operator's Manual 9

22.5 Changing RM Settings ........................................................................................................................................................................ 22-4

22.5.1 Accessing RM Menus ............................................................................................................................................................. 22-4

22.5.2 Setting the Apnea Alarm Delay .......................................................................................................................................... 22-5

22.5.3 Selecting TV or MV for Display ............................................................................................................................................ 22-5

22.5.4 Selecting Flow or Vol Waveform for Display ................................................................................................................... 22-5

22.5.5 Changing the Wave Sweep Speed..................................................................................................................................... 22-5

22.5.6 Changing the Wave Scale ..................................................................................................................................................... 22-6

22.5.7 Setting RR Source ................................................................................................................................................................... 22-6

22.6 Understanding the Respiratory Loops ........................................................................................................................................... 22-6

22.7 Zeroing the RM Module ..................................................................................................................................................................... 22-7

22.8 Calibrating the Flow Sensor .............................................................................................................................................................. 22-7

23 Monitoring NMT (from Mindray NMT Module) ........................................................................................................ 23-1

23.1 Introduction ........................................................................................................................................................................................... 23-1

23.2 Safety ....................................................................................................................................................................................................... 23-1

23.3 Stimulation Modes ............................................................................................................................................................................... 23-2

23.3.1 Train-Of-Four (TOF) ................................................................................................................................................................ 23-2

23.3.2 Single Twitch (ST).................................................................................................................................................................... 23-3

23.3.3 Post-Tetanic Count (PTC) ...................................................................................................................................................... 23-3

23.3.4 Double-Burst Stimulation (DBS) ......................................................................................................................................... 23-3

23.4 Preparing for NMT Measurement .................................................................................................................................................... 23-4

23.4.1 Skin Preparation ...................................................................................................................................................................... 23-4

23.4.2 Placing the Electrodes and Sensor .................................................................................................................................... 23-4

23.5 Accessing the NMT Setup Menu ...................................................................................................................................................... 23-5

23.6 Calibrating the NMT Measurement ................................................................................................................................................ 23-6

23.6.1 Starting/Stopping NMT Measurements ........................................................................................................................... 23-6

23.7 Change NMT Measurement Settings ............................................................................................................................................. 23-7

23.7.1 Changing Stimulation Mode ............................................................................................................................................... 23-7

23.7.2 Changing Stimulation Current ........................................................................................................................................... 23-7

23.7.3 Changing Pulse Width ........................................................................................................................................................... 23-7

23.7.4 Changing Measurement Interval ....................................................................................................................................... 23-7

23.8 Enabling Block Recovery Note ......................................................................................................................................................... 23-7

23.9 Adjusting Stimulation Tone Volume ............................................................................................................................................... 23-7

23.10 Understanding NMT Display .......................................................................................................................................................... 23-8

23.11 Recalling Calibration Information ................................................................................................................................................. 23-9

24 Device Integration .................................................................................................................................................... 24-1

24.1 Introduction ........................................................................................................................................................................................... 24-1

24.2 Safety Information ............................................................................................................................................................................... 24-1

24.3 Supported Devices and Device Connection Method ............................................................................................................... 24-2

24.4 Differences in Displayed Values ....................................................................................................................................................... 24-2

24.5 Devices Integrated Window .............................................................................................................................................................. 24-3

24.6 System Functions of the Monitor .................................................................................................................................................... 24-4

24.6.1 Alarms ......................................................................................................................................................................................... 24-4

24.6.2 Data Storage ............................................................................................................................................................................. 24-4

24.6.3 Recording and Printing ......................................................................................................................................................... 24-4

10 Passport 12m/Passport 17m Operator's Manual

25 Monitoring tcGas ...................................................................................................................................................... 25-1

25.1 Introduction ........................................................................................................................................................................................... 25-1

25.2 Safety Information ............................................................................................................................................................................... 25-1

25.3 Connecting an External Device ........................................................................................................................................................ 25-2

25.4 tcGas Parameters .................................................................................................................................................................................. 25-2

25.5 Displaying tcGas Parameters ............................................................................................................................................................ 25-3

25.6 Entering the tcGas Setup menu ....................................................................................................................................................... 25-3

25.7 Setting tcpCO

2

/tcpO

2

Unit ................................................................................................................................................................. 25-3

25.8 tcGas Display ......................................................................................................................................................................................... 25-3

26 Freezing Waveforms ................................................................................................................................................. 26-1

26.1 Freezing Waveforms ............................................................................................................................................................................ 26-1

26.2 Viewing Frozen Waveforms ............................................................................................................................................................... 26-1

26.3 Unfreezing Waveforms ....................................................................................................................................................................... 26-2

26.4 Recording Frozen Waveforms ........................................................................................................................................................... 26-2

27 Review ....................................................................................................................................................................... 27-1

27.1 Accessing Respective Review Windows ........................................................................................................................................ 27-1

27.2 Reviewing Graphic Trends ................................................................................................................................................................. 27-1

27.3 Reviewing Tabular Trends .................................................................................................................................................................. 27-2

27.4 Reviewing Events ................................................................................................................................................................................. 27-3

27.4.1 Marking Events ........................................................................................................................................................................ 27-3

27.4.2 Reviewing Events .................................................................................................................................................................... 27-4

27.5 Reviewing Waveforms......................................................................................................................................................................... 27-6

27.6 Reviewing OxyCRG .............................................................................................................................................................................. 27-7

28 Calculations ............................................................................................................................................................... 28-1

28.1 Introduction ........................................................................................................................................................................................... 28-1

28.2 Dose Calculations ................................................................................................................................................................................. 28-2

28.2.1 Performing Calculations ....................................................................................................................................................... 28-2

28.2.2 Selecting the Proper Drug Unit .......................................................................................................................................... 28-2

28.2.3 Titration Table .......................................................................................................................................................................... 28-3

28.2.4 Drug Calculation Formulas .................................................................................................................................................. 28-3

28.3 Oxygenation Calculations ................................................................................................................................................................. 28-3

28.3.1 Performing Calculations ....................................................................................................................................................... 28-3

28.3.2 Entered Parameters ................................................................................................................................................................ 28-4

28.3.3 Calculated Parameters and Formulas ............................................................................................................................... 28-4

28.4 Ventilation Calculations ...................................................................................................................................................................... 28-5

28.4.1 Performing Calculations ....................................................................................................................................................... 28-5

28.4.2 Entered Parameters ................................................................................................................................................................ 28-5

28.4.3 Calculated Parameters and Formulas ............................................................................................................................... 28-5

28.5 Hemodynamic Calculations .............................................................................................................................................................. 28-6

28.5.1 Performing Calculations ....................................................................................................................................................... 28-6

28.5.2 Entered Parameters ................................................................................................................................................................ 28-6

28.5.3 Calculated Parameters and Formulas ............................................................................................................................... 28-7

28.6 Renal Calculations ................................................................................................................................................................................ 28-7

Passport 12m/Passport 17m Operator's Manual 11

28.6.1 Performing Calculations ....................................................................................................................................................... 28-7

28.6.2 Entered Parameters ................................................................................................................................................................ 28-8

28.6.3 Calculated Parameters and Formulas ............................................................................................................................... 28-8

28.7 Understanding the Review Window .............................................................................................................................................. 28-8

29 Recording .................................................................................................................................................................. 29-1

29.1 Using a Recorder................................................................................................................................................................................... 29-1

29.2 Overview of Recording Types ........................................................................................................................................................... 29-1

29.3 Starting and Stopping Recordings ................................................................................................................................................. 29-2

29.4 Setting up the Recorder ..................................................................................................................................................................... 29-2

29.4.1 Accessing the Record Setup Menu .................................................................................................................................... 29-2

29.4.2 Selecting Waveforms for Recording .................................................................................................................................. 29-2

29.4.3 Setting the Realtime Recording Length .......................................................................................................................... 29-3

29.4.4 Setting the Interval between Timed Recordings .......................................................................................................... 29-3

29.4.5 Changing the Recording Speed ......................................................................................................................................... 29-3

29.4.6 Clearing Recording Tasks ...................................................................................................................................................... 29-3

29.4.7 Setting the IBP Waveform Overlap Recordings ............................................................................................................. 29-3

29.5 Loading Paper ....................................................................................................................................................................................... 29-3

29.6 Removing Paper Jam .......................................................................................................................................................................... 29-4

29.7 Cleaning the Recorder Printhead .................................................................................................................................................... 29-4

30 Printing ..................................................................................................................................................................... 30-1

30.1 Printer ...................................................................................................................................................................................................... 30-1

30.2 Connecting a printer ........................................................................................................................................................................... 30-1

30.3 Setting Up the Printer ......................................................................................................................................................................... 30-2

30.4 Starting Report Printouts ................................................................................................................................................................... 30-2

30.5 Stopping Report Printouts ................................................................................................................................................................ 30-3

30.6 Setting Up Reports .............................................................................................................................................................................. 30-3

30.6.1 Setting Up ECG Reports ........................................................................................................................................................ 30-3

30.6.2 Setting Up Tabular Trends Reports .................................................................................................................................... 30-3

30.6.3 Setting Up Graphic Trends Reports ................................................................................................................................... 30-3

30.6.4 Setting Up Realtime Reports ............................................................................................................................................... 30-4

30.7 End Case Reports .................................................................................................................................................................................. 30-4

30.8 Printer Statuses ..................................................................................................................................................................................... 30-4

30.8.1 Printer Out of Paper ............................................................................................................................................................... 30-4

30.8.2 Printer Status Messages ........................................................................................................................................................ 30-4

31 Other Functions ........................................................................................................................................................ 31-1

31.1 Analog Output....................................................................................................................................................................................... 31-1

31.2 Exporting the Log ................................................................................................................................................................................ 31-1

31.3 Transferring Data .................................................................................................................................................................................. 31-1

31.3.1 Data Export System ................................................................................................................................................................ 31-1

31.3.2 Transferring Data by Different Means ............................................................................................................................... 31-2

31.4 Nurse Call ................................................................................................................................................................................................ 31-2

31.5 iView System (for Passport 17m only) ............................................................................................................................................ 31-3

31.5.1 Start, Power off and Restart iView System ...................................................................................................................... 31-4

12 Passport 12m/Passport 17m Operator's Manual

31.5.2 Configuring Application Program ShortCuts ................................................................................................................. 31-4

31.5.3 Using PC Software .................................................................................................................................................................. 31-6

31.5.4 iView Window Close and Standby ..................................................................................................................................... 31-6

31.5.5 Recover iView System ............................................................................................................................................................ 31-7

31.5.6 Remote Login ........................................................................................................................................................................... 31-7

31.5.7 Using McAfee Solidifier ......................................................................................................................................................... 31-7

31.6 Network Setup ...................................................................................................................................................................................... 31-8

31.6.1 Setting the Network Type .................................................................................................................................................... 31-8

31.6.2 Wireless Network Connection (For Passport 12m Only) ............................................................................................. 31-8

31.6.3 Viewing the MAC Address ................................................................................................................................................. 31-10

31.6.4 Enabling the Data Encryption ........................................................................................................................................... 31-11

31.6.5 Connecting the Monitor to the CMS............................................................................................................................... 31-11

31.6.6 Setting the Network Service Quality Level .................................................................................................................. 31-12

31.6.7 Setting the Multicast Parameters.................................................................................................................................... 31-12

31.7 MLDAP .................................................................................................................................................................................................. 31-13

31.7.1 Setting MLDAP ..................................................................................................................................................................... 31-13

31.7.2 Selecting Password for User Authentication ............................................................................................................... 31-13

31.8 Using DVI-VGA Adapter Box ........................................................................................................................................................... 31-14

32 Batteries .................................................................................................................................................................... 32-1

32.1 Overview ................................................................................................................................................................................................. 32-1

32.2 Safety ....................................................................................................................................................................................................... 32-2

32.3 Battery Guidelines ................................................................................................................................................................................ 32-2

32.4 Installing or Replacing a Battery ...................................................................................................................................................... 32-3

32.5 Charging a Battery ............................................................................................................................................................................... 32-4

32.6 Conditioning a Battery ....................................................................................................................................................................... 32-4

32.7 Checking Battery Performance ........................................................................................................................................................ 32-5

32.8 Storing a Battery ................................................................................................................................................................................... 32-6

32.9 Recycling a Battery .............................................................................................................................................................................. 32-6

33 Care and Cleaning ..................................................................................................................................................... 33-1

33.1 General Points ....................................................................................................................................................................................... 33-1

33.2 Cleaning and Disinfecting the Main Unit/Module/SMR ........................................................................................................... 33-2

33.2.1 Approved Cleaning and Disinfecting Agents ................................................................................................................. 33-2

33.2.2 Cleaning the Main Unit/Module/SMR .............................................................................................................................. 33-3

33.2.3 Disinfecting the Main Unit/Module/SMR ........................................................................................................................ 33-3

33.3 Cleaning and Disinfecting the Accessories .................................................................................................................................. 33-4

33.3.1 Approved Accessories Cleaning and Disinfecting Agents .......................................................................................... 33-4

33.3.2 Cleaning the Accessories ...................................................................................................................................................... 33-7

33.3.3 Disinfecting the Accessories ................................................................................................................................................ 33-7

33.4 Sterilization ............................................................................................................................................................................................ 33-7

34 User Maintenance ..................................................................................................................................................... 34-1

34.1 General Inspection ............................................................................................................................................................................... 34-1

34.2 Maintenance and Testing Schedule................................................................................................................................................ 34-2

34.3 Checking Monitor and Module Information ................................................................................................................................ 34-3

Passport 12m/Passport 17m Operator's Manual 13

34.4 ECG Verification ..................................................................................................................................................................................... 34-3

34.5 NIBP Leakage Test................................................................................................................................................................................. 34-4

34.6 CO

2

Leakage Test .................................................................................................................................................................................. 34-5

34.7 AG Leakage Test .................................................................................................................................................................................... 34-5

34.8 Checking NMT Sensor ......................................................................................................................................................................... 34-5

34.9 Calibrating the Touchscreen ............................................................................................................................................................. 34-6

35 Accessories ................................................................................................................................................................ 35-1

35.1 ECG Accessories .................................................................................................................................................................................... 35-1

35.2 SpO

2

Accessories ................................................................................................................................................................................... 35-3

35.3 NIBP Accessories ................................................................................................................................................................................... 35-4

35.4 Temp Accessories .................................................................................................................................................................................. 35-5

35.5 IBP/ICP Accessories .............................................................................................................................................................................. 35-6

35.6 C.O. Accessories ..................................................................................................................................................................................... 35-6

35.7 CO

2

Accessories ..................................................................................................................................................................................... 35-7

35.8 AG Accessories ....................................................................................................................................................................................... 35-8

35.9 BIS Accessories ...................................................................................................................................................................................... 35-8

35.10 NMT Accessories (for Mindray NMT module) ............................................................................................................................ 35-8

35.11 RM Accessories .................................................................................................................................................................................... 35-8

35.12 CCO/SvO

2

Accessories ....................................................................................................................................................................... 35-8

35.13 ScvO

2

Accessories .............................................................................................................................................................................. 35-9

35.14 Installation Accessories .................................................................................................................................................................... 35-9

35.15 Micellaneous Accessories ............................................................................................................................................................. 35-10

36 Part Numbers for Canadian Use ............................................................................................................................... 36-1

A Product Specifications.................................................................................................................................................. A-1

A.1 Monitor Safety Specifications ............................................................................................................................................................... A-1

A.2 Physical Specifications ............................................................................................................................................................................ A-3

A.3 Hardware Specifications ........................................................................................................................................................................ A-4

A.4 Data Storage .............................................................................................................................................................................................. A-6

A.5 Wireless Network (For Passport 12m Only) ...................................................................................................................................... A-7

A.6 Measurement Specifications ................................................................................................................................................................ A-9

B EMC and Radio Regulatory Compliance ...................................................................................................................... B-1

B.1 EMC ............................................................................................................................................................................................................... B-1

B.2 Radio Regulatory Compliance .............................................................................................................................................................. B-5

C Default Configurations ................................................................................................................................................. C-1

C.1 Parameters Configuration ...................................................................................................................................................................... C-1

C.2 Routine Configuration .......................................................................................................................................................................... C-17

C.3 User Maintenance Items....................................................................................................................................................................... C-22

D Alarm Messages ............................................................................................................................................................ D-1

D.1 Physiological Alarm Messages ............................................................................................................................................................ D-1

D.2 Technical Alarm Messages .................................................................................................................................................................... D-3

14 Passport 12m/Passport 17m Operator's Manual

E SpO

2

Sensor Accuracy .................................................................................................................................................... E-1

E.1 Accuracy of Masimo SpO

2

Sensors ...................................................................................................................................................... E-1

E.2 Accuracy of Nellcor SpO

2

Sensors ...................................................................................................................................................... E3

F Symbols and Abbreviations ........................................................................................................................................... F-1

F.1 Symbols ........................................................................................................................................................................................................ F-1

F.2 Abbreviations ............................................................................................................................................................................................. F-2

Passport 12m/Passport 17m Operator's Manual 15

FOR YOUR NOTES

16 Passport 12m/Passport 17m Operator's Manual

1

Safety

1.1

Safety Information

WARNING

Indicates a potential hazard or unsafe practice that, if not avoided, could result in death or serious injury.

CAUTION

Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage.

NOTE

Provides application tips or other useful information to ensure that you get the most from your product.

1.1.1

Warnings

WARNING

Before putting the system into operation, the operator must verify that the equipment, connecting cables and accessories are in correct working order and operating condition.

To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth. If a properly grounded power outlet is not available, operate the monitor on battery power.

The equipment is not intended to be used within the Magnetic Resonance (MR) environment.

To avoid an explosion hazard, do not use the equipment in the presence of oxygen-rich atmospheres, flammable anesthetics, or other flammable agents (such as gasoline).

Before connecting the equipment to the power line, check that the voltage and frequency ratings of the power line are the same as those indicated on the equipment’s label or in this manual.

Do not open the equipment housings. All servicing or future upgrades must be carried out by the Mindray trained and authorized personnel.

Do not come into contact with patients during defibrillation. Otherwise serious injury or death could result.

Do not rely exclusively on the audible alarm system for patient monitoring. Adjustment of alarm volume to a low level or off may result in a hazard to the patient. Remember that alarm settings should be customized according to different patient situations and always keeping the patient under close surveillance is the most reliable way for safe patient monitoring.

Passport 12m/Passport 17m Operator's Manual 1-1

WARNING

PATIENTS WITH A PACEMAKER – On ventricular paced patients, episodes of Ventricular Tachycardia may not always be detected. Do not rely entirely upon the system’s automated arrhythmia detection algorithm.

Keep pacemaker patients under close surveillance.

The physiological data and alarm messages displayed on the equipment should be reviewed by a clinician before being used for diagnostic interpretation and treatment.

To avoid inadvertent disconnection, route all cables in a way to prevent a stumbling hazard. Wrap and secure excess cabling to reduce risk of entanglement by patients or personnel.

Do not touch the equipment’s metal parts or connectors when in contact with the patient; otherwise patient injury may result.

Never mix patient electrode types or brands. Dissimilar metals or other incompatibilities may cause considerable baseline drift and may increase trace recovery time after defibrillation.

To reduce the hazard of burns during high-frequency surgical procedure, ensure that the monitor’s cables and transducers never come into contact with the electro-surgery unit (ESU).

The neutral electrode of the electro-surgery unit (ESU) shall properly contact the patient. Otherwise, burns may result.

Magnetic and electrical fields are capable of interfering with the proper performance of the equipment. For this reason make sure that all external devices operated in the vicinity of the equipment comply with the relevant EMC requirements. Mobile phone and X-ray equipment are a possible source of interference as they may emit higher levels of electromagnetic radiation.

1.1.2

Cautions

CAUTIONS

Only use parts and accessories specified in this manual.

Disposable accessories are not designed to be reused. Reuse may cause a risk of contamination and affect the measurement accuracy.

Always install or carry the equipment properly to avoid damage caused by drop, impact, strong vibration or other mechanical force.

Dry the equipment immediately in case of rain or water spray.

When no battery is installed, ensure that the monitor is supplied with continuous electric power during operation. Sudden power failure may lead to data loss.

This equipment is intended for single patient use.

Store and use the equipment in specified environmental condition. The monitor and accessories may not meet the performance specification due to aging, stored or used outside the specified temperature and humidity range.

Disposable accessories are not designed to be reused. Reuse may cause a risk of contamination and affect the measurement accuracy.

Contact the Mindray service personnel for replacements if you find the housing is broken.

1-2 Passport 12m/Passport 17m Operator's Manual

1.1.3

Notes

NOTES

Put the equipment in a location where you can easily view and operate the equipment.

During normal use, the operator is expected to face the front of the equipment.

The equipment uses a mains plug as a means of isolation the mains power supply. Do not position the equipment in a place difficult to operate the mains plug.

Remove the battery before shipping the monitor or if it will not be used for an extended period of time.

Keep this manual in the vicinity of the equipment so that it can be conveniently referenced when needed.

The software was developed in compliance with IEC60601-1. The possibility of hazards arising from software errors is minimized.

This manual describes all features and options. Your equipment may not have all of them. They are not available in all geographies.

Remove the battery before transporting the equipment or if the equipment will not be used for a long time.

At the end of its service life, the equipment, as well as its accessories, must be disposed of in compliance with the guidelines regulating the disposal of such products. If you have any questions concerning disposal of the equipment, please contact Mindray.

When disposing of the packaging material, be sure to observe the applicable waste control regulations and keep it out of children’s reach.

1.2

Equipment Symbols

Some symbols may not appear on your equipment.

Refer to instruction manual/ booklet

Power ON/OFF (for a part of the equipment)

Alternating current

Battery indicator

Connector for satellite module rack

Video output Equipotentiality

USB connector iView network connector

Defibrillator

Check sensor

Measure/standby

Passport 12m/Passport 17m Operator's Manual

Network connector

Output

Zero key

Calibrate key

Inserted direction

1-3

Gas outlet Serial number

DATE OF MANUAFACTURE General warning sign

Electrostatic sensitive devices Plastics identification symbol

DEFIBRILLATION-PROOF TYPE CF

APPLIED PART

DEFIBRILLATION-PROOF TYPE BF

APPLIED PART

Pushing prohibited (wheels locked, no pushing)

Protected against vertically falling water drops per IEC 60529

The presence of this label indicates the machine was certified by ETL with the statement:

Conforms to AAMI Std. ES 60601-1, IEC Std. 60601-1-6, IEC Std. 60601-1-8, IEC Std. 60601-2-25, IEC

Std. 60601-2-26, IEC Std. 60601-2-27, IEC Std. 60601-2-34, IEC Std. 60601-2-49, IEC Std. 80601-2-30,

ISO Std. 80601-2-55, ISO Std. 80601-2-56, ISO Std. 80601-2-61, IEC Std.

60601-2-10

Certified to CSA Std. C22.2 NO. 60601-1, NO. 60601-6, NO. 60601-1-8, NO. 60601-2-25, NO.

60601-2-26,NO. 60601-2-27, NO. 60601-2-34, NO. 60601-2-49, NO. 80601-2-30, NO. 80601-2-55, NO.

80601-2-56, NO. 80601-2-61, NO.

60601-2-10

1-4 Passport 12m/Passport 17m Operator's Manual

2

The Basics

2.1

Monitor Description

2.1.1

Intended Use

The Passport 17m and Passport 12m patient monitors are intended for monitoring, displaying, reviewing, alarming, and transferring of multiple physiological parameters including ECG (3-lead , 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, QT analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO

2

), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (SvO

2

/ScvO

2

), carbon dioxide (CO

2

), Oxygen (O

2

), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS), respiration mechanics (RM), and neuromuscular transmission monitoring (NMT). The equipment also provides an interpretation of resting 12-lead ECG.

All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:

The arrhythmia detection, ST Segment analysis of Mortara algorithm, BIS, RM, CCO, SvO

2

/ScvO

2

, PAWP monitoring and NMT monitoring are intended for adult and pediatric patients only;

 ST Segment analysis of Mindray algorithm is intended for adult patients only;

 C.O. monitoring is restricted to adult patients only;

 ICG monitoring is only for use on adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.

The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

WARNING

This patient monitor is intended for use only by clinical professionals or under their guidance. It must only be used by persons who have received adequate training in its use. Anyone unauthorized or untrained must not perform any operation on it.

NOTE

Mindray algorithm is hereinafter referred as the Advanced ECG algorithm.

Passport 12m/Passport 17m Operator's Manual 2-1

2.1.2

Applied Parts

The applied parts of the Passport 12m/17m patient monitors are:

ECG electrodes and leadwires

SpO

2

sensor

 NIBP cuff

 Temp probes

 IBP transducer

 C.O. sensor

 CO

2

sampling line/nasal sampling cannula, and water trap

 AG sampling line, water trap, and airway adapter

BIS sensor

RM sensor

ScvO

2

sensor

NMT sensor and electrode

2.2

Main Unit

2.2.1

Front View

1

2

3 4 5

2-2

6

7 8

9

10

11

12

Passport 12m/Passport 17m Operator's Manual

1. Alarm lamp

When a physiological alarm or technical alarm occurs, this lamp flashes as defined below.

High level alarms:

Medium level alarms:

Low level alarms:

2. Display Screen the lamp quickly flashes red. the lamp slowly flashes yellow. the lamp lights yellow without flashing.

3. Power On/Off Switch

Pressing this switch turns the monitor on.

When the monitor is on, pressing and holding this switch turns the monitor off.

An indicator is built into this switch. It turns on when the monitor is on and turns off when the monitor is off.

4. AC power LED

It turns on when AC power is connected.

5. Battery LED

On: when at least one battery is installed in the Passport 12m monitor and the AC source is connected; when two batteries are installed in the Passport 17m monitor and the AC source is connected.

Off: when no battery is installed, only one battery is installed in the Passport 17m monitor, the installed battery is malfunctioning, or no AC source is connected when the monitor is powered off.

Flashing: when the monitor operates on battery power.

6. [

Alarm Reset

] button

Press to reset the alarms.

7. [

Alarm Pause

] button

Press to pause or restore alarms.

8. [

Freeze

] button

Press to freeze or unfreeze waveforms.

9. [

Record

] button

When a recording task is not started, pressing this button starts recording. When a recording task is in progress, pressing this button stops recording.

10. [

NIBP

] button

Press to start or stop NIBP measurements.

11. [

Menu

] button

If no menu is currently displayed on the screen, pressing it enters the main menu. If a menu is displayed on the screen, pressing it closes that menu.

12. Knob

Rotate the knob clockwise or counter-clockwise to move the cursor.

Press the knob to select one item, such as accessing a menu or confirming the selection.

Passport 12m/Passport 17m Operator's Manual 2-3

2.2.2

Side View

Passport 12m

1

1

2

1. Integral Module Racks

2. Contact

3. Compartment for CF storage card slot

4. Recorder

5. Battery compartment

Passport 17m

5

3

4

1

2

3

1

2-4

1. Integral Module Racks

2. Compartment for CF storage card slot

3. Recorder

4. Contact

4

Passport 12m/Passport 17m Operator's Manual

NOTE

To ensure good contact, clean the contacts regularly, as dust and dirt may collect on them. When cleaning the contacts, wipe them with cotton, dampened with alcohol. Using forceps is recommended.

2.2.3

Rear View

Passport 12m

Radiator

1

8

2

3

7

6

4

5 5

1.

AC Power Input

2.

Micro-D Connector: outputs analog ECG, IBP and defibrillator synchronization signals simultaneously, among which the ECG signals support pace pulses to be enhanced.

3.

Nurse Call Connector: connects the patient monitor to the hospital’s nurse call system through the nurse call cable

(

PN: 8000-21-10361

). Alarms are sent to nurses through the nurse call system, if configured to do so.

4.

Network Connector: a standard RJ45 connector which connects the monitor to the central monitoring system

(CMS) or other monitor for remote view through the LAN. It can also be used to transmit data.

5.

USB Connectors: connect such devices as the USB mice, USB keyboard, etc.

6.

SMR Connector: connects the satellite module rack (SMR) , the T1 docking station , or the N1 Dock .

Passport 12m/Passport 17m Operator's Manual 2-5

7.

Digital Video Interface (DVI): connects a secondary display, which extends the display capability of your monitor.

The contents displayed on the secondary display screen matches what displays on the monitor screen.

8.

Equipotential Grounding Terminal: when using the monitor together with other devices, connect their equipotential grounding terminals together to eliminate the potential differences between them.

Passport 17m

Radiator

2-6

1

2

3

4

11

10

5

6 7

8

9

1.

Equipotential Grounding Terminal: when using the monitor together with other devices, connect their equipotential grounding terminals together to eliminate the potential differences between them.

2.

USB Connectors: only used for iView maintenance and data transfer.

3.

SMR Connector: connects the satellite module rack (SMR) , the T1 docking station, or the N1 Dock.

4.

Digital Video Interface (DVI): connects a secondary display, which extends the display capability of your monitor.

The secondary display can be independently operated and controlled, and also display the contents different from the monitor screen.

5.

USB Connectors: connect the controlling devices (USB mouse and USB keyboard) of the secondary display.

Passport 12m/Passport 17m Operator's Manual

6.

iView Network Connector: a standard RJ45 connector which connects the iView system to external network.

7.

Network Connector: a standard RJ45 connector which connects the monitor to the central monitoring system

(CMS) or other monitor for remote view through the LAN. It can also be used to transfer data.

8.

Micro-D Connector: outputs analog ECG, IBP and defibrillator synchronization signals simultaneously, among which the ECG signals supports pace pulses to be enhanced.

9.

Nurse Call Connector: connects the monitor to the hospital’s nurse call system through the nurse call cable (

PN:

8000-21-10361

). Alarms are sent to nurses through the nurse call system, if configured to do so.

10. USB Connectors: connect such devices as the USB mouse, USB keyboard, etc.

11. AC Power Input

2.2.4

Bottom View (Passport 17m)

Battery compartment

Passport 12m/Passport 17m Operator's Manual 2-7

2.3

Satellite Module Rack

Two types of Satellite Module Racks (SMR) can be connected to the monitor through the SMR connector via a SMR cable.

One type is the OEM SMR and the other type is the Mindray SMR. Both of these SMRs provide 8 slots for mounting measurement modules. The number of modules mounted in the SMR depends on the types of modules, as different modules consume a different number of slots.

The following figures show the indicator and connectors on the OEM SMR.

Contact

Indicator

Handle

Slot

SMR connector

2-8 Passport 12m/Passport 17m Operator's Manual

The following figures show the indicator and connectors on the Mindray SMR.

Contact

Indicator

Slot

SMR connector: only the SMR 1 port is for use with the Passport

12m/17m monitors

As shown in the figures above, there is an indicator showing the status of the SMR:

On: when the SMR works normally.

Off: when the SMR disconnects from the monitor, there is a problem with the power, or the monitor shuts down.

NOTE

To ensure a good contact, clean the contacts regularly, as dust and dirt may collect on them. When cleaning the contacts, wipe them with cotton, dampened with alcohol. Using forceps is recommended.

Passport 12m/Passport 17m Operator's Manual 2-9

2.4

Modules

As shown below, the monitor supports the following modules:

SpO

2

module:

MPM:

IBP module:

C.O. module:

CCO/SvO

2

module:

CO

2

module:

AG module:

BIS module:

RM module:

Pulse oxygen saturation module.

Multi-parameter module. It can simultaneously monitor ECG, respiration, SpO

2

, temperature,

NIBP and IBP. Refer to section

2.4.1 Multi-Parameter Module

for more information.

Invasive blood pressure module.

Cardiac output module.

CCO/SvO

2

interface module, used to interface with Edwards Vigilance II

® monitor, Vigileo TM monitor, or EV1000 monitor.

Carbon dioxide module (including sidestream and microstream).

Anesthesia gas module. The functions of the O

2

modules can be incorporated into it.

Bispectral index module.

Respiration mechanics module.

ScvO

T1:

2 module:

BeneLink module:

Central venous oxygen saturation module.

Used as a multi-measurement module for monitoring ECG, respiration, SpO

2

, temperature,

NIBP and IBP. Refer to section

2.4.2 T1

for more information.

BeneLink module is used for transmitting information from a connected external device to the monitor.

Neuromuscular transmission module. NMT module:

N1: Used as a multi-measurement module for monitoring ECG, respiration, SpO

2

, temperature,

NIBP, IBP, and CO

2

. Refer to section

2.4.3

N1

for more information.

Under the maximum configuration, the monitor has one two-slot module rack, one three-slot module rack and one satellite module rack. The number of modules mounted in the monitor depends on the types of modules, as different modules consume a different number of slots.

You can plug-in and unplug modules during patient monitoring. To plug- in a module, insert the module until the lever on the module clicks into place and then push the lock key at the bottom in position to lock the module. To unplug a module, release the lock key, press the lever upwards and pull the module out.

2-10

Make sure that the indicator on the module light is on after the module is plugged in.

Passport 12m/Passport 17m Operator's Manual

2.4.1

Multi-Parameter Module

The multi-parameter module (MPM) incorporates multiple measurement modules. Two types of MPM modules can be connected to the Passport 17m or 12m monitor. These MPM modules are named as MPM 2.0 and MPM 3.0.

As shown below, the module name is located at the upper left corner, all hardkeys on the upper part, and all measurement connectors on the lower part. Other measurement modules look similar to the MPM.

Example MPM 2.0

2 3 4

1

5

6

7

8

6

9

10

11

Example MPM 3.0

12

Passport 12m/Passport 17m Operator's Manual 2-11

1.

Module name

2.

Setup key: press to enter the [

MPM Setup

] menu.

3.

Zero key: press to enter the [

Zero IBP

] menu.

4.

NIBP start/stop key: press to start or stop NIBP measurements.

5.

Indicator

On: when the monitor works correctly.

Flash: when the module is being initialized.

Off: when the module is either disconnected or broken.

6.

IBP cable connector

7.

ECG cable connector

8.

Temp probe 1 connector

9.

Temp probe 2 connector

10. NIBP Cuff connector

11. SpO

2

cable connector

12. Analog out connector: outputs defibrillation synchronization pulse, ECG, and IBP analog signal.

NOTE

MPM 3.0 is supported by the software version 05.33.00 and above.

MPM 3.0 is only available for the U.S.

2-12 Passport 12m/Passport 17m Operator's Manual

2.4.2

T1

T1 can be connected to the Passport 17m or 12m monitor as a multi-measurement module.

1

2

3

4

5

6

7

8

1.

Temp probe 1 connector

2.

Temp probe 2 connector

3.

IBP cable connector

4.

Multifunctional connector: outputs analog ECG, IBP and defibrillation synchronization signals.

5.

NIBP cuff connector

6.

ECG cable connector

7.

SpO

2

cable connector

8.

Speaker

When the T1 is disconnected from the Passport 12m or 17m, it can continue to monitor a patient as a stand-alone monitor running on battery power or external DC power supply. For details of using the T1 as a stand-alone monitor, refer to the

T1 Operating Manual

(PN: 046-006734-00).

NOTE

The Micro-D connector on the rear of the Passport 12m/17m monitor is disabled when the multifunctional connector of the T1 is in use.

Please do not charge more than one T1 simultaneously with either the internal module rack or the satellite module rack.

Passport 12m/Passport 17m Operator's Manual 2-13

2.4.3

N1

2.4.3.1

Overview

BeneVision N1 patient monitor (hereinafter abbreviated as N1) can be connected to the Passport 17m or 12m monitor as a multi-measurement module.

1 2 3

4

10

9

8

5

6

7

1.

Power switch

2.

Sample line connector of the sidestream CO

2

3.

Temperature probe connector

4.

Gas outlet

5.

ECG cable connector

6.

SpO

2

cable connector

7.

NIBP cuff connector

8.

Multifunctional connector: outputting analog ECG, IBP and defib synchronization signal s .

9.

Speaker

10.

IBP cable connector

When the N 1 is disconnected from the Passport 12m or 17m, it can continue to monitor a patient as a stand-alone monitor running on battery power or external DC power supply. For details of using the N 1 as a stand-alone monitor, refer to the

N 1 Operating Manual

(PN: 046-011405-00 ).

2-14

NOTE

The Micro-D connector on the rear of the Passport 12m/17m monitor is disabled when the multifunctional connector of the

N

1 is in use.

Please do not charge more than one

N

1 simultaneously with either the internal module rack or the satellite module rack.

The N1 is supported by the software version 05.33.00 and above.

The N1 is only available for the U.S.

Passport 12m/Passport 17m Operator's Manual

2.5

Display Screen

This monitor adopts a high-resolution TFT LCD to display patient parameters and waveforms. A typical display screen is shown below.

1

2

3

4

5

6

7

8

9

1.

Patient Information Area

This area shows the date, time, and the patient information such as department, bed number, patient name and patient category. indicates that no patient is admitted or the patient information is incomplete. If no patient is admitted, selecting this area enters the [Patient Setup] menu. For admitted patients, selecting this area enters the [Patient Demographics] menu.

2.

Alarm Symbols

 indicates alarms are paused.

 indicates alarms are reset. indicates alarm sounds are turned off.

 indicates the system is in alarm off status.

3.

Technical Alarm Area

This area shows technical alarm messages and prompt messages. When multiple messages occur, the messages scroll. Selecting this area shows the Technical Alarms list.

Passport 12m/Passport 17m Operator's Manual

2-1 5

2-1 6

4.

Physiological Alarm Area

This area shows physiological alarm messages. When multiple alarms occur, the messages scroll. Select this area and the physiological alarm list displays.

5.

Waveform Area

This area shows measurement waveforms. The waveform name displays at the upper left corner of the waveform.

Select this area and the corresponding measurement setup menu displays.

6.

Parameter Area A

This area shows measurement parameters. Each monitored parameter has a parameter window and the parameter name displays at the upper left corner. The corresponding waveform of each parameter is displayed in the same row in the waveform area. Select this area and the corresponding measurement setup menu displays.

7.

Parameter Area B

For the parameter values displayed in this area, their corresponding waveforms are not displayed.

8.

Prompt Message Area

This area shows the prompt messages, network status icons, battery status icons, etc.

For Passport 12m, this area shows the currently selected CMS if the [Select CMS] function is enabled. If no CMS is

selected, this area displays “???”. Refer to section

0 Selecting a CMS

for detail.

For details about battery status symbols, refer to chapter

32 Batteries.

 indicates the monitor is successfully connected to a wired network.

 indicates the monitor has failed to connect to a wired network.

 indicates the wireless function (optional) is connected. indicates the wireless function (optional) is not connected.

 indicates a CF storage card is inserted. indicates a USB drive is inserted.

indicates a secondary display is connected.

[Screen Setup] button

9.

QuickKeys Area

This area contains QuickKeys that provide quick access to functions.

Passport 12m/Passport 17m Operator's Manual

2.6

QuickKeys

A QuickKey is a configurable graphical key, located at the bottom of the main screen. They give you fast access to functions. Their availability, and the order in which they appear on your screen, depend on your monitor configuration.

To configure QuickKeys:

1.

Select [

Main Menu

]

[

Maintenance

>>]

[

Manage Configuration

>>]

→ enter the required password

[

Ok

].

2.

In the [

Manage Configuration

] menu, select [

Edit Config.

>>].

3.

In the pop-up menu, select the desired configuration and then select [

Edit

].

4.

In the pop-up menu, select [

Screen Setup

>>].

5.

Select the [

Select QuickKeys

] tab, and then configure your desired QuickKeys and the order of them.

The following QuickKeys are available:

Scroll left to display more QuickKeys.

[

Main Menu

] QuickKey

Enter the [Main Menu]

[

Alarm Setup

] QuickKey

Enter the [Alarm Setup] menu

[

NIBP Measure

] QuickKey

Enter the [NIBP Measure] menu

[

Zero IBP

] QuickKey

Enter the [Zero IBP] menu

[

Alarm Pause

] QuickKey

Pause or restore alarms

[

Patient Setup

] QuickKey

Enter the [Patient Setup] menu

[

Realtime Print

] QuickKey

Start the realtime print

[

Minitrends

] QuickKey

Have a split-screen view of minitrends

[

Load Config

.] QuickKey

Enter the [

Load configuration

] menu

[

Start C.O.

] QuickKey

Enter the C.O. measurement window

[

Calculations

] QuickKey

Enter the [

Calculations

] menu

[

7-Lead ECG

] QuickKey

Enter the 7-lead ECG full screen

Scroll right to display more QuickKeys.

[

Standby

] QuickKey

Enter standby mode

[

Review

] QuickKey

Enter the [Review] menu

[

Stop All

] QuickKey

Stop all NIBP measurement

[

Alarm Reset

] QuickKey

Reset the alarm system

[

Screens

] QuickKey

Enter the [Screens] menu

[

Manual Event

] QuickKey

Trigger a manual event

[

Print Setup

] QuickKey

Enter the [Print Setup] menu

[

Volume Setup

] QuickKey

Enter the [Volume Setup] menu

[

iView

] QuickKey (For Passport 17m only)

Have the iView

[

Loops

] QuickKey

View respiratory loops

[

Others

] QuickKey

Enter the [View Other Patient] window

[

12-Lead ECG

] QuickKey

Enter the 12-lead ECG full screen

Passport 12m/Passport 17m Operator's Manual 2-1 7

[

Parameters

] QuickKey

Enter the [Parameters] menu

[

NIBP STAT

] QuickKey

Start NIBP STAT measurement

[

Night Mode

] QuickKey

Enter the [Night Mode] menu

[

Privacy Mode

] QuickKey

Enter the privacy mode

[

Discharge Patient

] QuickKey

Enter the [

Discharge Patient

] menu

[

OxyCRG

] QuickKey

Enter the [oxyCRG] window

[

PAWP

] QuickKey

Enter the PAWP measurement screen

[

Unit Setup

] QuickKey

Enter the [Unit Setup] menu

[

CPB Mode

] QuickKey

Enter the CPB mode

NOTE

There is space for only 12 user configurable Quickkeys on the Passport 17m, and 10 user configurable

Quickkeys on the Passport 12m.

2-1 8

Passport 12m/Passport 17m Operator's Manual

3

Basic Operations

3.1

Installation

WARNING

Connect only approved devices to this equipment. Devices connected to the equipment must meet the requirements of the applicable IEC standards (e.g. IEC 60950 safety standards for information technology equipment and IEC 60601-1 safety standards for medical electrical equipment). The system configuration must meet the requirements of the IEC 60601-1 medical electrical systems standard. Any personnel who connect devices to the equipment’s signal input/output port are responsible for providing evidence that the safety certification of the devices has been performed in accordance to the IEC 60601-1. If you have any questions, please contact Mindray.

If it is not evident from the equipment specifications whether a particular combination with other devices is hazardous, for example, due to summation of leakage currents, please consult the manufacturer or an expert in the field. A determination must be made that the proposed combination will not negatively affect the devices themselves or the patient's safety.

Use only screws and brackets specified by Mindray, or the screw could touch the internal battery, causing monitor damage.

CAUTION

The equipment should be installed by authorized Mindray personnel.

NOTE

The software equipment copyright is solely owned by Mindray. No organization or individual shall resort to modifying, copying, or exchanging it or to any other infringement on it in any form or by any means without due permission.

Many settings in the patient monitor are password protected. It is recommended to change the default passwords and keep the passwords safe. Passwords should only be changed by authorized personnel.

Contact your department manager or biomedical engineering department for the passwords used at your facility.

3.1.1

Unpacking and Checking

Before unpacking, examine the packing case carefully for signs of damage. If any damage is detected, contact the carrier or Mindray.

If the packing case is intact, open the package and remove the equipment and accessories carefully. Check all materials against the packing list and check for any mechanical damage. Contact Mindray in case of any problem.

Passport 12m/Passport 17m Operator's Manual 3-1

CAUTION

When disposing of the packaging material, be sure to observe the applicable waste control regulations and keep it out of children’s reach.

Before use, please verify whether the packages are intact, especially the packages of single use accessories.

In case of any damage, do not apply it to patients.

NOTE

Save the packing case and packaging material as they can be used if the equipment must be reshipped.

3.1.2

Environmental Requirements

The equipment operating environment must meet the requirements specified in this manual.

The equipment operating environment should be reasonably free from noises, vibration, dust, corrosive, flammable and explosive substances. To maintain good ventilation, the equipment should be at least 2 inches (5 cm) away from surrounding objects.

When the equipment is moved from one place to another, condensation may occur as a result of temperature or humidity difference. In this case, never start the system before the condensation disappears.

CAUTION

Make sure that the equipment operating environment meets the specific requirements. Otherwise unexpected consequences, e.g. damage to the equipment, could result.

3.2

Getting Started

3.2.1

Connecting to Power Source

Using AC Power Source

To use the AC power source, connect one end of the power cord with the AC power input on the equipment’s back panel and the other end with a wall AC mains outlet.

WARNING

Always use the accompanying power cord with the monitor.

Use the battery if the integrity of the protective earth conductor or the protective earthing system in the installation is in doubt.

3-2 Passport 12m/Passport 17m Operator's Manual

Using a Battery

This monitor can be equipped with rechargeable lithium-ion battery. If a battery is installed, the monitor system automatically switches to the battery for power if AC power is interrupted.

NOTE

When a battery has been stored for a long time, or the battery is depleted, recharge the battery at once.

Otherwise, the low battery may not be sufficient to power the monitor if the AC power is unavailable.

Refer to chapter

32 Batteries

for details.

Connecting Accessories

Insert the connector of accessory cables necessary for the measurements to be performed by your monitor to the

corresponding connector on the monitor. For details about the parameter connectors, refer to section

2.4 Modules

.

3.2.2

Turning Power On

Once the monitor is installed, before beginning measurements:

1. Check the monitor, SMR and plug-in modules for any mechanical damage, and make sure that all external cables, plug-ins and accessories are properly connected.

2. Check the power supply specification is met if mains power is used. Only use a power outlet that is properly grounded.

3. Plug the power cord into the AC power source. If you run the monitor on battery power, ensure that the battery is sufficiently charged.

4. Press the power on/off switch on the monitor’s front panel.

The monitor will perform alarm system self-test during start-up. After pressing the power on/off button, the system sounds a beep and the alarm lamp simultaneously turns yellow, then red, and then turns off, followed by the start-up screen being shown. Then the start-up screen disappears. The alarm system self-test succeeds. The monitor enters the normal monitoring screen.

WARNING

Do not use the monitor on a patient if you suspect it is not working properly, or if it is mechanically damaged. Contact your service personnel or Mindray.

NOTE

Carefully check if the system performs the self-test as described above. Contact your service personnel or

Mindray if the self-test is abnormal.

Passport 12m/Passport 17m Operator's Manual 3-3

3.2.3

Starting Monitoring

To start monitoring:

1. Decide which measurements you want to take.

2. Connect the required modules, patient cables and sensors.

3. Make sure that the patient cables and sensors are correctly connected.

4. Make sure that the patient settings, such as [

Patient Cat.

], [

Paced

], etc, are appropriate for your patient.

5. Refer to the appropriate measurement section for details of how to perform the measurements you require.

3.3

Turning Off the Monitor

Before turning off the monitor:

1. Ensure that the monitoring of the patient has been completed.

2. Disconnect the cables and sensors from the patient.

3. Make sure to save or clear the patient monitoring data as required.

4. Press and hold the power on/off switch for more than 2 seconds to turn off the monitor. “

System is shutting down…

” is displayed on the screen and then the monitor shuts down.

CAUTION

Press and hold the power on/off switch for 10 seconds to forcibly shut down the monitor if it could not be shut down normally. This may cause loss of patient data.

The monitor restores the latest configuration if it restarts within 60 seconds after a power failure. The monitor restores the default configuration, rather than the latest configuration, if it restarts 120 seconds after a power failure. The monitor may load either the latest configuration or the default configuration if it restarts from 60-120 seconds after a power failure.

Power failure may cause data corruption on the SD card. It is recommended to turn off the monitor according to the normal procedures. Do not unplug the power cord unless charged battery/batteries are installed. To operate on DC power, the Passport 12m requires one battery while the Passport 17m requires two batteries.

NOTE

To completely disconnect the power source, unplug the power cord and then remove the battery/batteries.

3-4 Passport 12m/Passport 17m Operator's Manual

3.4

Using the Knob

 Rotate the knob clockwise or counterclockwise to move the cursor.

Press the knob to select one item, such as accessing a menu or confirming the selection.

3.5

Using a Mouse

You can use the USB mouse supplied with the equipment as a monitor input device. The USB mouse can be plugged and unplugged with the monitor turned on.

When you are using a mouse:

By default, the left mouse-button is the primary button and the right one the secondary button.

Clicking the primary button is equal to pressing the knob or selecting the touchscreen.

 The secondary button is disabled.

To define the right mouse-button as the primary button:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2. Select [

Others >>

] to enter the [

Others

] menu.

3. Select [

Primary Button

] and then select [

Right

] from the pop-up list.

3.6

Using Keys

The monitor has three types of keys:

Softkey: A softkey is a graphic key on the screen, giving you fast access to certain menus or functions. The monitor has two types of softkeys:

Parameter keys: Each parameter area can be seen as a softkey. You can enter a parameter setup menu by selecting its corresponding parameter or waveform area.

QuickKeys: QuickKeys are configurable graphical keys, located at the bottom of the main screen. For details, refer to the section

2.6

QuickKeys.

Hardkeys: A hardkey is a physical key on a monitoring device, such as the [

Main Menu

] key on the monitor’s front panel.

 Pop-up Keys: Pop-up keys are task-related keys that appear automatically on the monitor screen when needed. For example, the Confirm pop-up key appears only when you need to confirm a change.

Passport 12m/Passport 17m Operator's Manual 3-5

3.7

Using the On-screen Keyboard

The on-screen keyboard enables you to enter information.

 Use the key to delete the previously entered character.

Use the key to toggle between uppercase and lowercase letters.

 Select

 Select

to confirm what you have entered and close the on-screen keyboard.

to access the symbol keyboard.

Select to exit the symbol keyboard.

3.8

Using the Touchscreen

Select screen items by pressing them directly on the patient monitor’s screen. You can enable or disable touchscreen operation by pressing and holding the [

Main Menu

] QuickKey for 3 seconds. A padlock symbol touchscreen operation is disabled.

displays if

3.9

Using the [Main Menu]

To enter the [

Main Menu

], select the on-screen QuickKey or the [

Menu

] hardkey on the monitor’s front panel. Most of monitor operations and settings can be performed through the [

Main Menu

].

3-6

1

2

3

4

Passport 12m/Passport 17m Operator's Manual

4.

Other menus are similar to the [

Main Menu

] and contain the following parts:

1. Heading: describes the title for the current menu.

2. Main body: displays options, buttons, prompt messages, etc. The menu button with “>>’’ open a secondary window to reveal more options or information.

3. Online help area: displays help information for the highlighted menu item.

: select to exit the current menu. If changes are made in the menu, they take effect after selecting this button.

3.10

Setting the Screen

You can enter the [

Screen Setup

] window as shown below by selecting the [

Screen Setup

] icon in the prompt message area. In this window, you can arrange the position of the parameters and waveforms. Unselected parameters or waveforms will not display.

Area C

Area A

Area B

The ECG parameter and the first ECG waveform always display in the first row. The configurable areas can be classified as

Area A, Area B, and Area C.

 In Area A, you can choose to display the parameter windows and their waveforms (if one exists). Each parameter and the associated waveform are displayed in the same row.

In Area B, you can choose to display the parameter windows. But if all parameters in area C are set to [

Off

], both the parameter and waveform selected in the first row of area B will be displayed.

In Area C, you can choose to display the timer and any parameters without associated waveforms.

The screen automatically adjusts to ensure the best view based on your screen setup.

Passport 12m/Passport 17m Operator's Manual 3-7

If no corresponding parameter or waveform is displayed after the module is inserted, perform the following inspections:

Check the connection between the module and lead, cable, sensor, or external device.

Check whether the [

The display setup for XX is disabled

] message is displayed in the prompt message area and the [

Screen Setup

] icon desired display configuration.

is flashing. If yes, select this icon to enter the [

Screen Setup

] window for the

 Check that the parameter is turned on in [

Parameters Switch

] window.

CAUTION

Unallocated parameters in the [Screen Setup] window will not display. However, the monitor still sounds alarms for these parameters.

3.11

Displaying the Timer

To display the timer in the main screen:

1. Select the [

Screens

] button in prompt message area to access the [

Screens

] window.

2. Select [

Screen Setup

] tab.

3. In the Area C, select [

Timer

] from the drop-down list of the desired parameter area. Refer to section

3.10 Setting the Screen

for Area C.

4. Select to exit the window. The main screen will display the timer.

Select [

Start

] or [

Pause

] to start or pause timing.

Select [

Clear

] to clear current timing result.

Select [

Setup

] to access the [

Timer Setup

] window, in which you can set the [

Direction

] to [

Up

] or [

Down

]. If you select [

Down

], you should set:

[

Run Time(h:min:s)

]: The available time range is 0 to 100 hours, and the default time is 5 minutes.

[

Reminder Vol

]: During the last 10 seconds of the countdown, the system issues reminder tone. The available volume range is 0 to 10. 0 means off, and 10 the maximum volume.

3-8 Passport 12m/Passport 17m Operator's Manual

3.12

Changing General Settings

This chapter covers only general settings such as language, brightness, date and time, etc. Refer to measurement and other settings in their respective sections.

3.12.1

Setting up a Monitor

To install a monitor or change its location:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2. In the [

User Maintenance

] menu, select, [

Monitor Name

], [

Department

] and [

Bed No

.] or edit their settings.

3.12.2

Changing Language

To change the language:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2. In the [

User Maintenance

] menu, select [

Language

] and then select the desired language.

3. Restart the monitor.

NOTE

The changed language is applied only after the monitor is restarted.

Passport 12m/Passport 17m Operator's Manual 3-9

3.12.3

Adjusting the Screen Brightness

To adjust the screen brightness:

1. Select the [

Main Menu

]

[

Screen Setup >>

]

[

Brightness

].

2. Select the appropriate setting for the screen brightness. 10 is the brightest, and 1 is the dimmest.

If the monitor operates on battery power, setting a dimmer screen brightness will prolong the battery operating time.

When the monitor enters Standby mode, the screen automatically changes to the dimmest setting.

3.12.4

Showing/Hiding the Help Text

The monitor provides online help information. The user can display or hide the help text as needed.

1. Select [

Main Menu

]

[

Screen Setup >>

].

2. Select [

Help

] and select either [

On

] or [

Off

].

3.12.5

Setting the Date and Time

To set the date and time:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

System Time >>

].

2. Set the date and time.

3. Select [

Date Format

] and then select [

yyyy-mm-dd

], [

mm-dd-yyyy

] or [

dd-mm-yyyy

].

4. Select [

Time Format

] and then select [

24h

] or [

12h

].

If your monitor is connected to a central station (CMS), the date and time are automatically synchronized to that CMS. In that case, you cannot change the date and time settings on your monitor.

CAUTION

Changing date and time affects the storage of trends and events and may cause data loss. Save or record any needed data prior to changing the date and time.

3.12.6

Adjusting Volume

Alarm Volume

To adjust the alarm volume:

1. Select the [

Volume Setup

] QuickKey, or [

Main Menu

]

[

Alarm Setup >>

].

2. Input the password as required. See section

31.7.2 Selecting Password for User Authentication.

3. Select [

Others

] tab.

3. Select [

Alm Volume

] and then select the appropriate volume: X-10, in which X is the minimum volume, depending on the set minimum alarm volume (refer to

section

7.4.1

Setting the Minimum Alarm Volume

), and 10 is the

maximum volume.

3-10 Passport 12m/Passport 17m Operator's Manual

Key Volume

When you press the navigation knob, the touchscreen, or the hardkeys on the panel, the monitor prompts you by making a sound of the key volume you have set.

To adjust the key volume:

1. Select the [

Volume Setup

] QuickKey, or [

Main Menu

]

[

Screen Setup >>

].

2. Select [

Key Volume

] and then select the appropriate volume. 0 means off, and 10 is the maximum volume.

QRS Volume

The QRS tone is derived from either the HR or PR, depending on which is currently selected as the alarm source in [

ECG

Setup

] or [

SpO

2

Setup

]. When monitoring SpO

2

, there is a variable pitch tone which changes as the patient’s saturation level changes. The tone pitch rises as the saturation level increases and falls as the saturation level decreases. The tone volume is user adjustable.

To adjust the QRS volume:

1. Select the [

Volume Setup

] QuickKey, or the ECG parameter window

[

Others >>

], or the SpO

2

parameter window.

2. Select [

QRS Volume

] or [

Beat Vol

] and then select the appropriate volume. 0 means off, and 10 is the maximum volume.

3.13

Setting Parameters

3.13.1

Switching the Parameters On/Off

To switch the parameters on or off, select [

Main Menu

]

[

Screen Setup >>

]

[

Screen Layout >>

]

[

Parameters

Switch

], or [

Screens

] QuickKey

[

Parameters Switch

]. When a parameter is switched off, its corresponding parameter module stops working, and its parameter value and waveform are not shown on the monitor display.

NOTE

ECG is always selected, and you can not switch it off.

Passport 12m/Passport 17m Operator's Manual 3-11

3.13.2

Accessing the Parameters Menu

Select [

Parameters >>

] from the [

Main Menu

] or select the [

Parameters

] QuickKey at the bottom of the screen to enter the [

Parameters

] menu where you can access each parameter’s setup menu. You can further select [

Module Status >>

] to enter the menu as shown below. Your display may be configured to look slightly different depending on the modules mounted.

This menu displays the measurement modules mounted in the two-slot module rack, three-slot module rack and satellite module rack from top to bottom. Beside each measurement connector is the measurement label. The color in which a measurement connector appears matches the status as follows:

(colored) indicates that the module is turned on.

(grey) indicates that the module is turned off.

indicates a module name conflict.

indicates a module error.

3.13.3

Removing a Module Conflict

Generally, the monitor can support only one type of parameter modules simultaneously. Otherwise, the module conflict message will be prompted. For example, if a CO

2 module is already loaded and then another CO

2 module is inserted, your monitor will display a module conflict. To use one module, just pull the other module out.

But for IBP measurements, it can support three independent IBP modules and the IBP module on the MPM simultaneously, while for SpO

2

monitoring, it can support one independent SpO

2

module and the SpO

2

module on the

MPM simultaneously.

3-12 Passport 12m/Passport 17m Operator's Manual

3.14

Using a CF Storage Card

A CF storage card is used to save patient data such as trend data, waveform data, etc., during patient monitoring. In case of a sudden power failure, the patient data can be retrieved from the CF storage card after the monitor restarts.

To insert a CF storage card, open the compartment and then insert the card until the button pops out.

To remove the CF storage card:

1. In the [

Main Menu

], select [

Unload Storage Card

], or [

Patient Data

]

[

Unload Storage Card

]. You can also click

icon in the lower right corner of the screen.

2. Select [

Ok

] from the pop-up menu to unload the CF storage card. A status message shown in the prompt message area will report completion of the unloading.

3. Press the button until the CF storage card pops out.

To browse the data saved in the CF storage card:

1. Select [

Main Menu

]

[

Patient Data >>

]

[

Historical Data >>

].

2. Select a patient whose data you want to view from the [

Patient Data List

] and then select [

Review

].

3. Select [

Data Review

].

As reviewing the patient’s data history is just like reviewing the patient’s current data, you can refer to the chapter

27

Review

for details.

CAUTION

Use only the CF storage card specified by Mindray.

Please take measures against the static electricity by wearing a wrist strap when you remove the CF card.

NOTE

If no CF storage card is used, all the data saved to the internal memory will be lost in case of monitor shut-down or sudden power interrupt.

Unload the CF storage card before removing it from the monitor. Otherwise it may cause damage to the data in the card.

The system deletes the oldest data stored in the CF card automatically when the remaining space is insufficient.

Passport 12m/Passport 17m Operator's Manual 3-13

3.15

Operating Modes

Your monitor has different operating modes. Some are password protected. This section lists the major operating modes.

3.15.1

Monitoring Mode

This is the normal, everyday working mode that you use for monitoring patients. Your monitor automatically enters the

Monitoring mode after being turned on.

3.15.2

Night Mode

To avoid disturbing the patient, Night mode may be used.

To activate the Night mode:

1. Select the [

Night Mode

] QuickKey or [

Main Menu

]

[

Screen Setup >>

]

[

Night Mode >>

].

2. In the pop-up menu, set the desired brightness, alarm volume, QRS volume, key volume, NIBP end tone, or whether to stop NIBP measurements or not. When [

Stop NIBP

] is selected, all the NIBP measurements terminate after entering the Night mode.

3. Select the [

Enter Night Mode

] button.

To cancel the Night mode:

1. Select the [

Night Mode

] QuickKey or [

Main Menu

]

[

Screen Setup >>

]

[

Night Mode >>

].

2. Select [

Ok

] in the pop-up.

CAUTION

Before entering Night mode, confirm the brightness settings, alarm volume, QRS volume, and key volume.

Consider the potential risk when the setting value is low.

3-14 Passport 12m/Passport 17m Operator's Manual

3.15.3

Privacy Mode

Privacy mode is only available when an admitted patient at a monitor is also monitored by the CMS.

To activate the Privacy mode, select [

Main Menu

]

[

Screen Setup >>

]

[

Privacy Mode

].

The monitor does the following after activating Privacy mode:

The screen turns blank and the message [

Under monitoring. Press any key to exit the privacy mode.

] displays.

Monitoring and data storing continue but patient data is only visible at the CMS.

Alarms can still be triggered. But all audible alarms are suppressed and the alarm light is deactivated at the monitor.

All system sounds are suppressed, including heart beat tone, pulse tone, all prompt tones, etc.

To cancel the privacy mode, press any key.

The monitor exits the Privacy mode automatically in any of the following situations:

 The monitor disconnects from the CMS.

 The alarm [

Battery Too Low

] or [

System will shut down soon. Please replace the batteries or use the external power.

] is presented.

The touchscreen is locked automatically in the Privacy mode.

WARNING

In Privacy mode, all audible alarms are suppressed and the alarm light is deactivated at the monitor. Alarms sound only at the CMS.

NOTE

The Privacy mode can be used only when the monitor is connected to a CMS.

3.15.4

Standby Mode

In Standby mode, you can temporarily stop patient monitoring without turning off the monitor. To enter the Standby mode, select the [

Standby

] QuickKey .

Passport 12m/Passport 17m Operator's Manual 3-15

FOR YOUR NOTES

3-16 Passport 12m/Passport 17m Operator's Manual

4

Managing Patients

4.1

Admitting a Patient

The monitor displays and stores physiological data in trends as soon as a patient is connected. This allows you to monitor a patient that is not admitted yet. However, it is recommended that you admit a patient so that you can clearly identify them on recordings, reports and networking devices.

To admit a patient:

1. Select the [

Patient Setup

] QuickKey, or [

Main Menu

]

[

Patient Setup >>

].

2. Select [

Admit Patient

].

If a patient has previously been admitted, a message [

Are you sure to discharge the current patient and admit a new patient?

] pops up. Then select [

Ok

] to clear any previous patient data. If you do not erase data from the previous patient, the new patient’s data is combined with the previous patient’s data. The monitor makes no distinction between the old the new patient data.

If no patient has been previously admitted, you can choose either:

[

Yes

] to apply the data saved in the monitor to the new patient, or

[

No

] to clear the data saved in the monitor.

3. In the [

Patient Demographics

] menu, enter the demographic details, of which:

[

Patient Cat.

] allows the user to select patient size (Neo, Ped, Adu) which will determine the way your monitor processes and calculates some measurements, and what safety and alarm limits are applied for your patient.

[

Paced

] determines whether to show pace pulse marks on the ECG waveform. When the [

Paced

] is set to [

No

], pace pulse marks are not shown in the ECG waveform.

4. Select [

Ok

].

WARNING

Make sure the [Patient Cat.] and [Paced] fields always contain a value, regardless of whether the patient is fully admitted or not.

For paced patients, you must set [Paced] to [Yes]. If it is incorrectly set to [No], the monitor could mistake a pace pulse for a QRS and fail to alarm when the ECG signal is too weak.

For non-paced patients, you must set [Paced] to [No].

For details about patient paced status, refer to section

8.3.4 Checking Paced Status

.

Passport 12m/Passport 17m Operator's Manual 4-1

4.2

Quick Admitting a Patient

Use [

Quick Admit

] only if you do not have the time or information to fully admit a patient. Complete the rest of the patient demographic details later. Otherwise, the symbol always displays in the patient information area.

1. Select the [

Patient Setup

] QuickKey, or [

Main Menu]

[

Patient Setup >>

].

2. Select [

Quick Admit

]. If a patient has been admitted, select [

Ok

] to discharge the current patient. If no patient is admitted, you can choose either:

[

Yes

] to apply the data in your monitor to the new patient, or

[

No

] to clear any previous patient data

.

3. Enter the Patient Category and Paced Status for the new patient, and then select [

Ok

].

4.3

Setting the Monitor Location

To set the monitor location, follow this procedure:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2. Input the following location of the monitor:

[

Facility

]: your facility name.

[

Department

]: your department name.

[

Room No.

]: room number.

[

Bed No.

]: bed number.

4.4

Querying and Obtaining Patient Information

The monitor can obtain patient information from Hospital Information System (hereafter called HIS) through the eGateway. To query or obtain patient information from the HIS:

1. Select [

Main Menu]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password

[

Network

Setup >>

]

[

Gateway Comm Setup >>

], and set [

IP Address

] and [

Port

]. Set [

ADT Query

] to [

On

]. Then select

[

Ok

].

2. Select the patient information area to enter the [

Patient Demographics

] menu.

3. Select [

Obtain Patient Info. >>

] to enter the [

Obtain Patient Information

] menu.

4. Input a query condition and then select [

Query

]. The monitor displays the requested patient information.

5. Select a patient and then click [

Import

] to update the corresponding patient information.

6. Select to exit the [

Obtain Patient Information

] menu.

NOTE

4-2

The option [Obtain Patient Information] is available in the [Patient Setup] menu only when [ADT Query] is set to [On].

When obtaining patient information from the HIS, only patient information is updated on the monitor. The patient’s physiological data is not changed and the patient is not discharged.

Passport 12m/Passport 17m Operator's Manual

4.5

Querying from Local Facility

You can query the patient information from either the local facility or all networked facilities. To set where to query, follow this procedure:

1. Select [

Main Menu]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password

[

Network

Setup >>

]

[

Gateway Comm Setup >>

].

2. Set [

Query From Local Facility

].

Select [

Yes

] to query only from local facility.

Select [

No

] to query from all networked facilities.

4.6

Associating Patient Information

After associating patient information with the HIS, the monitor automatically updates the patient information if corresponding HIS information changes. The monitor can associate the patient’s MRN (Medical Record Number), visit number, first name, last name, date of birth, and gender with the HIS.

NOTE

A keyword takes effect only when being defined in eGateway. Refer to

eGateway Integration Manager

Installation Guide

(PN: 046-002447-00) for details.

The monitor displays corresponding patient information only when all the keywords have been entered.

4.7

Editing Patient Information

To edit the patient information after a patient has been admitted, when the patient information is incomplete, or when you want to change the patient information:

1. Select the [

Patient Setup

] QuickKey, or [

Main Menu

]

[

Patient Setup >>

].

2. Select [

Patient Demographics

] and then make the required changes.

3. Select [

Ok

].

You can also input the patient’s visit number in the [

Patient Demographics

] menu, but the [

Visit Number

] option needs to be enabled.

To display the [

Visit Number

] option in the [

Patient Demographics

] menu:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2. Set [

Visit Number

] to [

On >>

].

Passport 12m/Passport 17m Operator's Manual 4-3

4.8

Discharging a Patient

To discharge a patient:

1.

Select the [

Dicharge Patient

] QuickKey, or [

Main Menu

]

[

Patient Setup >>

]

[

Discharge Patient

].

2.

In the pop-up menu, you can either:

Select [

Ok

] to discharge the current patient, or

Select [

Standby

] then [

Ok

]. The monitor enters the standby mode after discharging the current patient, or

Select [

Cancel

] to exit without discharging the patient.

If you need to print the end case report, refer to section

30.7

End Case Reports

for instructions.

NOTE

After discharging an admitted patient, you can review the historical data of this patient by selecting [Main

Menu]

[Patient Data

>>

]

[Historical Data

>>

]. When the CF card is full, the oldest patient data will be deleted automatically. Therefore, you cannot always view all the historical data.

4.9

Transferring Patient Data

You can transfer a patient with an MPM, T1, or N1 to a new location without re-entering the patient demographic information or changing the settings. Transferring of patient data enables you to review the patient’s condition history.

The patient data that can be transferred includes: patient demographics, trend data, alarm events and parameter alarm limits. You can also use a USB drive or a CF storage card to transfer data between two monitors.

From the [

User Maintenance

] menu, select [

Others >>

]. In the pop-up menu, you can set [

Transferred Data Length

].

The default is [

4 h

]. You can also set [

Data Transfer Method

]. The default is [

Off

].

WARNING

Do not discharge a patient before the patient is successfully transferred.

After a patient is successfully transferred, check if the patient settings (especially patient category, paced status and alarm limits settings, etc) on the monitor are appropriate for this patient.

Only when [Apply Module Settings] is set to [On] and select [Continue Patient in MPM] or [Continue Patient in

Transfer Module

], can the IBP labels be transferred along with the MPM/T1

/N1

.

NOTE

The system automatically enables the HR alarm and lethal arrhythmia alarm after transferring the patient data.

The N1 is only available for the U.S.

4-4 Passport 12m/Passport 17m Operator's Manual

4.9.1

Transferring Patient Data via MPM/T1/N1

Familiarizing yourself with the data respectively stored in the monitor, T1, N1, or MPM helps you understand the effects incurred by transferring patients with an MPM/T1/N1.

Contents stored

Patient demographics

(Name, Bed No., Gender, etc.)

Trend data

Data

Calculation data

(Dose calculations, oxygenation calculations, etc.)

Settings

Event data

(Marked events, alarm events, etc.)

Monitor settings

(Alarm pause, alarm volume, etc.)

Parameter settings

(Alarm limits, etc.)

In the patient monitor In the MPM

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

No

No

No

Yes

In the T1

Yes

Yes

No

Yes

No

Yes

In the N1

Yes

Yes

No

Yes

No

Yes

Before transferring a patient with an MPM/T1/N1:

1.

Select [

Main Menu

]

[

Maintenance

]

[

User Maintenance >>

]

→ enter the required password.

2.

Select [

Others >>

].

3.

Set [

Data Transfer Method

] to [

Module

].

4.

Set [

Apply Module Settings

] to [

On

]. If your monitor does not have this option, the system applies the

MPM/T1/N1’s settings by default.

5.

Set [

Data Transfer Strategy

] to:

[

Always Ask

]: when an MPM/T1/N1 is inserted to a monitor, the monitor displays a pop-up window asking the operator to choose what kind of data the monitor will apply.

[

Continue Module

]: when an MPM/T1/N1 is inserted to a monitor, the monitor will apply the data saved in the MPM/T1 /N1 .

[

Continue Monitor

]: when an MPM/T1/N1 is inserted to a monitor, the monitor will apply the data saved in the monitor.

To transfer the patient:

1.

Disconnect MPM/T1/N1 from the original monitor.

2.

Connect MPM/T1/N1 to the destination monitor.

3.

If there is a mismatch between the MPM/T1/N1 and monitor, the system will automatically display the [

Select

Patient

] menu, from which you can choose the data set you want to continue using for this patient, either:

[

Continue Patient in Monitor

]: continue with the patient data and settings in the monitor, deleting all patient data and settings in MPM/T1/N1 and copying all data in the monitor to MPM/T1/N1.

[

Continue Patient in MPM

] or [

Continue Patient in Transfer Module

]: continue with the patient data and settings in MPM/transfer module. The transer module refers to T1 and N1. Discharge the patient in the monitor.

The monitor then automatically admits the patient and copies all data from MPM/T1/N1.

Passport 12m/Passport 17m Operator's Manual 4-5

[

New Patient

]: select this button if none of the information is correct. This deletes all data in the monitor and

MPM/T1/N1 and lets you admit a new patient on the monitor. In this case, you need to re-enter the patient demographics. The monitor will restore the settings according to the patient category.

[

Same Patient

]: select this button if the patient demographics are different, but it is the same patient. This merges the patient’s trend data in the monitor and MPM/T1/N1 and copies the settings in MPM/T1/N1 to the monitor also. It only appears when the critical patient information of an identical patient ID is different between the monitor and the MPM/T1/N1 module.

4.

Select [

Ok

].

Operations

Continue Patient in Monitor

Continue Patient in MPM/Continue

Patient in Transfer Module

New Patient

Same Patient

Examples of applications

1. Replace MPM/T1/N1 during patient monitoring.

2. After the patient is admitted, connect the MPM/T1/N1.

A patient is monitored using MPM/T1/N1. You need to transfer the patient, e.g. from a ward (original monitor) to the operating room (destination monitor).

Connect the MPM/T1/N1 before admitting a new patient. However, the monitor and/or MPM/T1/N1 store the previous patient’s data and settings.

A patient is admitted by a monitor, to which MPM/T1/N1 used in another monitor for monitoring this patient is connected.

4.9.2

Transferring Patient Data via a CF Storage Card or USB Drive

4.9.2.1

Transferring Data from the Monitor to a Storage Medium

To transfer data from the monitor to a storage medium:

1.

Insert a storage medium into the monitor. Select [

Main Menu

]

[

Patient Setup >>

].

2.

Select [

Transfer to Storage Medium

]. In the pop-up menu, you can:

Select [

Ok

] to transfer the patient data, or

Select [

Cancel

] to exit the menu.

3.

Wait until the following message appears: [

Transfer to storage medium successful. Please remove the CF storage card.

] or [

Transfer to storage medium successful. Please remove the USB drive.

].

4.

Remove the CF storage card or USB drive from monitor.

4.9.2.2

Transferring Data from the Storage Medium to the Monitor

To transfer data from a storage medium to the monitor:

1.

Connect the storage medium to the destination monitor.

2.

Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password

[Others >>]

→ [

Data Transfer Method

] → [

USB Drive

] or [

Storage Card

]. Then:

Select [

Transfer

] to transfer the patient data to the monitor, or

Select [

Cancel Transfer

] to cancel the transfer operation.

Select [

Unload Storage Card

] or [

Unload USB Drive

] to unload the card or USB drive without transferring the patient data.

4-6 Passport 12m/Passport 17m Operator's Manual

3.

After you select [

Transfer

] in the pop-up menu, you can select the patient data contents that need to be transferred. [

Patient Demographics

] must be selected. After selecting [

Ok

], the monitor compares the patient information stored in both the storage medium and monitor and manages the patient data based on the following:

Different Patient: The monitor erases all the current patient data, transfers the patient data from the storage medium, and loads the configuration according to the patient category.

Same Patient: In the pop-up dialog box, you can:

Select [

Yes

] to merge the patient data in the monitor and storage medium.

Select [

No

] to erase all the current patient data in the monitor and to transfer the patient data from the storage medium.

4.

Wait until the message [

Transfer from storage medium successful.

] appears before removing the storage medium.

CAUTION

The USB drive you use may be write-protected. In this case, make sure the USB drive is in read/write mode.

Do not remove the storage medium during data transfer process. Otherwise, data files may be damaged.

The normal monitoring function will be affected during data export. Do not perform any monitoring activity during data export.

4.10

Connecting to a CMS

The monitor supports connection to a CMS. If your monitor is connected to a CMS:

All patient information, measurement data and settings on the monitor can be transferred to the CMS.

 All patient information, measurement data and settings can be displayed simultaneously on the monitor and CMS.

For some functions such as editing patient information, admitting a patient, discharging a patient, starting/stopping NIBP measurements, etc., bi-directional control can be achieved between your monitor and the

CMS.

For details, refer to

BeneVision Central Monitoring System Operator’s Manual (PN: 046-007960-00 and PN:

046-010879-00)

.

Passport 12m/Passport 17m Operator's Manual 4-7

FOR YOUR NOTES

4-8 Passport 12m/Passport 17m Operator's Manual

5

Managing Configurations

5.1

Introduction

When performing continuous patient monitoring, the clinical professional often needs to adjust the monitor’s settings according to the patient’s condition. The collection of all these settings is called a configuration. Allowing you to configure the monitor more efficiently, the monitor provides different sets of configurations to accommodate the varying patient categories and departments. You can change the settings from a default configuration and then save it as a user configuration.

The default configurations provided for your monitor are department-oriented. You can choose either from:

General

OR

ICU

NICU

CCU

Each department has three different sets of configurations one for each patient category [

Neo

], [

Ped

], and [

Adu

]. For the

configuration items and their default values and user maintenance items, see Appendix

C Default Configurations

.

NOTE

The configuration management function is password protected. The configuration management tasks must be performed by authorized personnel. Please contact your department manager or biomedical engineering department for the password used at your facility.

Passport 12m/Passport 17m Operator's Manual 5-1

5.2

Accessing the [Manage Configuration] Menu

To access the [

Manage Configuraion

] menu:

1. Press the [

Menu

] hardkey on the monitor’s front panel to enter the [

Main Menu

].

2. Select [

Maintenance >>

]

[

Manage Configuration >>

]. Enter the required password and then select [

Ok

].

5.3

Changing Department

If the current department configuration is not the one you want to view, you can select [

Change Department >>

] in the

[

Manage Configuration

] menu and then choose the one you want for viewing as shown below.

5-2

NOTE

Changing the department will delete all current user configurations.

Passport 12m/Passport 17m Operator's Manual

5.4

Setting Default Configuration

The monitor will load the pre-set default configuration in the following cases:

 The monitor restarts after being switched off for more than 120 seconds.

A patient is admitted.

A patient is discharged.

Patient data is cleared.

Patient category is changed.

To set the default configuration:

1. Select [

Select Default Config. >>

] in the [

Manage Configuration

] menu.

2. In the [

Select Default Config.

] menu, select [

Load the Latest Config.

] or [

Load Specified Config.

].

When you select [

Load Specified Config.

], the restored configuration is subject to the patient category (adult, pediatric or neonate). This configuration can be either factory configuration or a saved user configuration. As an example, select

[

Default Adu Config.

] and then select [

Defaults

] or user configuration(s).

When you select [

Load the Latest Config

], the latest configuration is loaded when the monitor is started or a patient is admitted.

NOTE

To identify which configuration is restored when the monitor starts, enter the main screen to check the prompt information at the lower part of the screen.

5.5

Saving Current Settings

Current settings can be saved as a user configuration. Up to 5 user configurations can be saved.

To save current settings:

1. Select [

Save Current Settings As >>

] in the [

Manage Configuration

] menu.

2. In the pop-up dialog box, enter the configuration name and then select [

Ok

]. The current settings are saved as a user configuration with the name you entered.

Passport 12m/Passport 17m Operator's Manual 5-3

5.6

Editing Configurations

To edit an existing configuration:

1. Select [

Edit Config. >>

] in the [

Manage Configuration

] menu. The following menu appears.

2. The pop-up menu shows the existing monitor configurations. Selecting [

Config. on USB drive >>

] will show the existing configurations on the USB drive. Select the desired configuration and then select the [

Edit

] button. The following menu appears.

5-4

3. Select [

Alarm Setup >>

], [

Screen Setup >>

] or [

Parameters >>

] to enter the corresponding menu in which settings can be changed. The changed alarm setup items will be marked in red.

4. You can select [

Save

] or [

Save as

] to save the changed configuration. Select [

Save

] to overwrite the original configuration. Select [

Save as

] to save the changed configuration as another file with another name.

Passport 12m/Passport 17m Operator's Manual

5.7

Deleting a Configuration

To delete a configuration:

1. Select [

Delete Config. >>

] in the [

Manage Configuration

] menu. The pop-up menu shows the existing user configurations on the monitor.

2. Select [

Config. on USB drive >>

] to show the existing user configurations on the USB drive. Select the user configurations you want to delete and then select [

Delete

].

3. Select [

Yes

] in the pop-up.

5.8

Transferring a Configuration

When installing several monitors with identical user configurations, it is not necessary to set each unit separately. Use a

USB drive to transfer the configuration from monitor to monitor.

To export the current monitor’s configuration:

1. Connect the USB drive to the monitor’s USB port.

2. Select [

Export Config. >>

] in the [

Manage Configuration

] menu.

In the [

Export Config

.] menu, select the configurations and [

User Maintenance Settings

] to export. Then select the [

Export

] button. A status message will report completion of the transfer.

To import the configuration from the USB drive to the monitor:

1. Connect the USB drive to the monitor’s USB port.

2. Select [

Import Config. >>

] in the [

Manage Configuration

] menu.

3. In the [

Import Config.

] menu, select the configurations and [

User Maintenance Settings

] to import. Then select the [

Import

] button. A status message will report completion of the transfer.

5.9

Loading a Configuration

You may make changes to some settings during operation. However, these changes or the pre-selected configuration may not be appropriate for the newly admitted patient. Therefore, the monitor allows you to load a desired configuration to ensure that all the settings are appropriate for your patient.

To load a configuration,

1. Select [

Load Configuration >>

] from the [

Main Menu

]. The pop-up menu shows the existing configurations on the monitor.

2. Select [

Config. on USB drive >>

] to show the existing configurations on the USB drive.

3. Select a desired configuration.

4. Select [

View

] to see the configuration details. In the pop-up menu, you can select [

Alarm Setup >>

], [

Screen Setup

>>

] or [

Parameter >>

] to view the corresponding contents. The alarm setup items which are different than those currently used are marked in red.

5. Select [

Load

].

Passport 12m/Passport 17m Operator's Manual 5-5

NOTE

The monitor may configure some settings by default when you load a configuration of different software version with the current configuration.

5.10

Restoring the Latest Configuration Automatically

During operation, you may make changes to some settings. However, these changes may not be saved as a user configuration. To prevent the changes from being lost in case of a sudden power failure, the monitor stores the configuration in real time. The saved configuration is the latest configuration.

The monitor restores the latest configuration if it restarts within 60 seconds after the power failure. And it will restore the default configuration rather than the latest configuration if it restarts 120 seconds after the power failure. The monitor loads either the latest configuration or the default configuration if it restarts from 60-120 seconds after the power failure.

5-6 Passport 12m/Passport 17m Operator's Manual

6

User Screens

6.1

Configuring Your Screens

You can configure your monitor’s screens by setting:

Waveform sweep mode

Waveform line size

The color in which each measurement’s numerics and waveforms are displayed

The parameter to monitor

CAUTION

Changing some settings may be hazardous. Therefore, those settings are password-protected and can be modified by authorized personnel only. Once a change is made, notify those who use the monitor.

6.1.1

Changing the Waveform Line Size

To change the waveform line size:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2. Select [

Others >>

].

3. Select [

Wave Line

] and select [

Thick

], [

Med

] or [

Thin

].

6.1.2

Changing Measurement Colors

To change the measurement colors:

1. Select [

Main Menu

]

[

Screen Setup >>

]

[

Measurement Color Setup >>

].

2. Select the color box next to your desired parameter and then select a color from the pop-up menu.

6.1.3

Changing Screen Layout

Select the [

Screens

] QuickKey, or [

Main Menu

]

[

Screen Setup >>

]

[

Screen Layout >>

] to enter the [

Screens

] menu.

You can choose the desired screen type in the [

Choose Screen

] window.

You can select the parameters and waveforms you want to view in the [

Screen Setup

] window. For details, please

refer to the section

3.10 Setting the Screen

.

 You can select the parameters you want to view on the big numerics screen in the [

Big Numerics Screen Setup

] window.

You can switch the connected parameter modules on or off in the [

Parameters Switch

] window. If a parameter module is switched off, parameter values and waveforms will not display on the screen.

Passport 12m/Passport 17m Operator's Manual 6-1

6.1.4

Setting the Waveform Sweep Mode

To set the waveform sweep mode:

1. Select [

Main Menu

]

[

Screen Setup >>

].

2. Select [

Sweep Mode

] and then select [

Refresh

] or [

Scroll

].

[

Refresh

]: The waveforms are refreshed from left to right.

[

Scroll

]: The waveforms move from the right to the left with time passing by.

6.2

Viewing Minitrends

6.2.1

Having a Split-Screen View of Minitrends

You can split the normal screen so that the left hand side continuously shows graphic minitrends beside waveforms as shown in the figure below.

To have a split-screen view of minitrends:

Select the [

Minitrends

] QuickKey, or

Select the [

Screens

] QuickKey

[

Choose Screen

]

[

Minitrends Screen

]

, or

Select [

Main Menu

]→

[

Screen Setup >>

]→[

Screen Layout >>

]→[

Choose Screen

]

[

Minitrends Screen

]

.

6-2

Minitrend View

Passport 12m/Passport 17m Operator's Manual

The split-screen view provides minitrends for multiple parameters. In each field, the label and scale are respectively displayed at the top and left. The time is displayed at the bottom of the minitrends view as shown below.

Time

6.2.2

Setting Minitrends

Select the minitrends area. From the pop-up [

Minitrend Setup

] menu:

Select the parameters to be displayed, or

Select [

Minitrend Length

] and then select the appropriate setting.

6.3

Viewing OxyCRG

To have a split screen view of OxyCRG:

Select the [

OxyCRG

] QuickKey, or

Select the [

Screens

] QuickKey

[

Choose Screen

]

[

OxyCRG Screen

]

, or

Select [

Main Menu

]→

[

Screen Setup >>

]→[

Screen Layout >>

]→

[

Choose Screen

]

[

OxyCRG Screen

]

.

1

2

Passport 12m/Passport 17m Operator's Manual

3

4

5

6

6-3

The split-screen view covers the lower part of the waveform area and shows HR trend, SpO

2

trend, SpO

2 b trend, RR trend and a compressed waveform (CO

2

or Resp wave). At the bottom, there are controls:

1. Event

You can enter the [

Review

] menu by selecting the [

Event

] button.

2. Trend length list box

In the trend length list box, you can select [

1 min

], [

2 min

], [

4 min

], or [

8 min

].

3. Setup

Select the [

Setup

] button to enter the [

Setup

] menu, in which you can select the parameters for display, the time length to be saved before and after an event, and the scale of the graphic trends and waveform. The trend area can display two parameter trends, e.g. HR trend and RR trend, simultaneously.

4. Auto Scale

Select the [

Auto Scale

] button, and the system automatically adjusts the scaling.

5. Print

Select [

Print

] to print the real-time OxyCRG to a printer.

6. Record

Select [

Record

] to print the currently displayed OxyCRG trends to the recorder.

6.4

Viewing Other Patients

6.4.1

Care Group

The Passport 12m can support a Care Group of up to 10 monitors and the Passport 17m can support a Care Group of up to 16 monitors.This lets you:

View information on the monitor screen from another bed in the same Care Group.

Be notified of physiological and technical alarm conditions at the other beds in the same Care Group.

To have a Care Group:

1. Open the [

View Other Patient

] window by:

Selecting [

Others

] QuickKey, or

Selecting [

Screens

] QuickKey

[

Choose Screen

]

[

View Others Screen

]→

, or

Selecting [

Main Menu

]→

[

Screen Setup >>

]→[

Screen Layout >>

]→

[

Choose Screen

]

[

View Others

Screen

]→

.

2. Select [

Setup

] in the [

View Other Patient

] window.

3. Select the desired monitors from the [

Connected Monitor List

], and then select the button. The selected monitors constitute a Care Group.

6-4 Passport 12m/Passport 17m Operator's Manual

NOTE

Monitors using software version prior to 05.17.00 cannot view or be viewed by monitors of software version

05.17.00 or later.

Re-set the Care Group if the monitor is moved to a department or different LAN.

6.4.2

Viewing the Care Group Overview Bar

The Care Group overview bar is located at the bottom of the [

View Other Patient

] window. In the overview bar, the department and bed label for any Care Group beds are displayed. For telemetry, # is displayed before the department label. The color in which a Care Group bed appears matches its status:

Red: indicates the bed is having high-level physiological alarms or the telemetry is in alarm, such as nurse call or event.

Yellow: indicates the bed is having medium-level or low-level physiological alarms, or medium-level technical alarms.

Blue: indicates the bed is having low-level technical alarms.

Light grey: indicates the bed fails to be networked.

Dark grey: indicates the bed is in the Standby mode.

You can view a Care Group bed’s alarms by selecting it from the Care Group, and you can select the [

View This Patient

] button to view the bed in the [

View Other Patient

] window.

For more details about Care Group alarms, refer to the Chapter

7 Alarms

.

Passport 12m/Passport 17m Operator's Manual 6-5

6.4.3

Understanding the View Other Patient Window

When you first open the [

View Other Patient

] window, the monitor automatically selects a device from the network to display in the [

View Other Patient

] window.

4

1

5

2

3

The [

View Other Patient

] window covers the lower part of the waveform area and consists of:

1. Information Area: shows the patient information (including department, bed number, patient name, etc.), and network status symbol.

2. View Area: shows physiological waveforms and parameters. You can switch a waveform area to a parameter area by selecting your desired waveform area and then selecting [

Switch to Parameter Area

], or switch a parameter area to a waveform area by selecting your desired parameter area and then selecting [

Switch to Waveform Area

].

3. Care Group Overview Bar. See section

6.4.2 Viewing the Care Group Overview Bar

for more information.

4. Message Area: shows physiological, technical and prompt messages from the currently viewed patient monitor. It also shows the alarm given by the device such as nurse call or event. By selecting this area, you can enter the

[

Alarm Information List

] to view all physiological, technical and prompt messages coming from the currently viewed patient.

5. [

Alarm Reset

] button: resets alarms from the currently viewed bed. Refer to section

7.11.3 Resetting Care Group

Alarms

.

When the [

Reset Other Bed’s Alarms

] is disabled, no button will appear here.

6-6 Passport 12m/Passport 17m Operator's Manual

Additionally, you can change a waveform or parameter for viewing:

To change a waveform for viewing, select the waveform segment where you want a new waveform to appear and then select the waveform you want from the pop-up menu.

To change a parameter for viewing, select the parameter window where you want a new parameter to appear and then select the parameter you want from the pop-up menu.

WARNING

The data presented in the [View Other Patient] window has a delay. Do not rely on this window for realtime data.

The icon indicates disconnection to the viewed monitor. You cannot view the monitor when this icon is displayed.

6.5

Understanding the Big Numerics Screen

To enter the big numerics screen:

1. Select the [

Screens

] QuickKey, or [

Main Menu

]

[

Screen Setup >>

]

[

Screen Layout >>

].

2. Select [

Big Numerics

]

.

You can select your desired parameters to display in this screen: select the [

Screens

] QuickKey

[

Big Numerics Screen

Setup

] and then select the parameters you want. For parameters having a waveform, the waveform will also be displayed.

Passport 12m/Passport 17m Operator's Manual 6-7

FOR YOUR NOTES

6-8 Passport 12m/Passport 17m Operator's Manual

7

Alarms

Alarms, triggered by an abnormal vital sign or technical issue with the monitor, as visually and audibly indicated to the user.

WARNING

A potential hazard can exist if different alarm presets are used for the same or similar equipment in any single area, e.g. an intensive care unit or cardiac operating room.

If your monitor is connected to a CMS, alarms can be controlled remotely. Remote suspension, inhibition, and reset of monitor alarms via the CMS may cause a potential hazard. For details, refer to the CMS’s operator’s manual.

For troubleshooting specific alarms, see appendix

D Alarm Messages

.

7.1

Alarm Categories

The monitor’s alarms can be classified into two categories: physiological alarms and technical alarms.

1. Physiological alarms

Physiological alarms, also called patient status alarms, are triggered by a monitored parameter value that violates set alarm limits or an abnormal patient condition. Physiological alarm messages are displayed in the physiological alarm area.

2. Technical alarms

Technical alarms, also called system status alarms, are triggered by a device malfunction or a patient data distortion due to improper operation or mechanical problems. Technical alarm messages are displayed in the technical alarm area.

Apart from the physiological and technical alarm messages, the monitor will show some messages telling the system status or patient status. Messages of this kind are included in the prompt message category and usually displayed in the prompt information area. Some prompt messages that indicate the arrhythmia events are displayed in the physiological alarm area. For some measurements, their related prompt messages are displayed in their respective parameter windows.

Passport 12m/Passport 17m Operator's Manual 7-1

7.2

Alarm Levels

The monitor’s alarms can be classified into three severity categories: high level, medium level and low level.

High level

Medium level

Low level

Physiological alarms

Indicate that the patient is in a life threatening situation, such as Asystole,

Vfib/Vtac and so forth, and an emergency treatment is required.

Indicate that the patient’s vital signs appear abnormal and immediate treatment is required.

Indicate that the patient’s vital signs appear abnormal and immediate treatment may be required.

Technical alarms

Indicate a severe device malfunction or an improper operation which could make it possible that the monitor cannot detect critical patient status and thus threaten the patient’s life.

Indicate a device malfunction or an improper operation, which may not threaten the patient’s life but may compromise the monitoring of vital physiological parameters.

Indicate a device malfunction or an improper operation which may compromise a certain monitoring function but will not threaten the patient’s life.

7.3

Alarm Indicators

When an alarm occurs, the monitor will indicate it to the user through visual or audible alarm indications.

 Alarm lamp

 Alarm message

 Flashing numerics

Audible alarm tones

7.3.1

Alarm Lamp

If a technical alarm or physiological alarm occurs, the alarm lamp will flash. The color and flashing frequency match the alarm level as follows:

High level alarms:

Medium level alarms:

Low level alarms: the lamp quickly flashes red the lamp slowly flashes yellow the lamp lights yellow without flashing

7.3.2

Alarm Messages

When an alarm occurs, an alarm message will appear in the technical or physiological alarm area. For physiological alarms, the asterisk symbols (*) before the alarm message match the alarm level as follows:

High level alarms:

Medium level alarms:

Low level alarms:

***

**

*

Additionally, the alarm message has different background colors to match the alarm levels:

 High level alarms:

 Medium level alarms:

Low level alarms: red yellow yellow

7-2 Passport 12m/Passport 17m Operator's Manual

You can view the alarm messages by selecting the physiological or technical alarm area. Refer to section

2.5

Display

Screen

for details.

7.3.3

Flashing Numeric

If an alarm triggered by an alarm limit violation occurs, the numeric of the measurement in alarm will flash every second, and the corresponding alarm limit will also flash at the same frequency indicating the high or low alarm limit is violated.

7.3.4

Audible Alarm Tones

The alarm tone is distinct from heart beat tone, keystroke tone and pulse tone in frequency. This monitor has three alarm tones and patterns: ISO, Mode 1 and Mode 2. For each pattern, the alarm tones identify the alarm levels as follows:

ISO pattern:

High level alarms: triple+double+triple+double beep

 Mode 1:

High level alarms: high-pitched single beep

Medium level alarms: double beep

Medium level alarms: triple beep

Low level alarms: single beep

Low level alarms:

Mode 2: low-pitched single beep

High level alarms: high-pitched triple beep

Medium level alarms: double beep

Low level alarms: low-pitched single beep

NOTE

When multiple alarms of different levels occur simultaneously, the monitor will select the highest level alarm, light the alarm lamp and give alarm sounds accordingly, while all the alarm messages are displayed circularly on the screen.

Some physiological alarms, such Asystole, are exclusive. They have identical alarm tones and alarm lights with normal high level physiological alarms, but their alarm messages are displayed exclusively. That is, when an exclusive physiological alarm and a normal high level physiological alarm are triggered simultaneously, only exclusive physiological alarm message will be displayed.

Passport 12m/Passport 17m Operator's Manual 7-3

7.3.5

Alarm Status Symbols

Apart from the aforementioned alarm indicators, the monitor still uses the following symbols telling the alarm status:

 indicates alarms are paused. indicates alarm is reset.

 indicates the alarm sound is turned off.

 indicates individual measurement alarms are turned off or the system is in alarm off status.

7.4

Alarm Tone Configuration

7.4.1

Setting the Minimum Alarm Volume

To set the minimum alarm volume:

1.

Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2.

Select [

Alarm Setup >>

] to enter the [

Alarm Setup

] menu.

3.

Select [

Minimum Alarm Volume

] and then select a value between 0 and 10.

The minimum alarm volume refers to the minimum value you can set for the alarm volume, which is not affected by user or factory default configurations. The minimum alarm volume setting remains unchanged when the monitor shuts down and restarts.

7.4.2

Changing the Alarm Volume

To change the alarm volume:

1.

Select the [

Volume Setup

] QuickKey or the [

Alarm Setup

] QuickKey

[

Others

], or [

Main Menu

]

[

Alarm

Setup >>

]

→ enter the required password

[

Others

].

2.

Select the appropriate volume from [

Alm Volume

]: X-10, in which X is the minimum volume, depending on the set minimum alarm volume, and 10 is the maximum volume.

3.

Select [

High Alarm Volume

] to set the volume of the high priority alarm as [

Alm Volume+0

], [

Alm Volume+1

] or

[

Alm Volume+2

].

4.

Select [

Reminder Vol

] to set the volume of the reminder tone as

[High

], [

Med

] or [

Low

].

When alarm volume is set to 0, the alarm sound is turned off and a screen. symbol appears on the alarm symbols area of the

7-4 Passport 12m/Passport 17m Operator's Manual

7.4.3

Setting the Interval Between Alarm Sounds

If you choose the ISO pattern, you can change the interval between alarm tones. To change the interval between alarm tones:

1.

Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2.

Select [

Alarm Setup >>

] to enter the [

Alarm Setup

] menu.

3.

Select [

High Alarm Interval (s)

], [

Med Alarm Interval (s)

] and [

Low Alarm Interval (s)

] in order and then select the appropriate settings.

You cannot change the interval between alarm tones if you choose mode 1 or 2 as your desired alarm tone pattern. For these two patterns, the interval between alarm tones identifies the alarm levels as follows:

Mode 1:

Interval between high level alarm tones:

Interval between medium level alarm tones:

Interval between low level alarm tones: continuously

5 s

20 s

Mode 2:

Interval between high level alarm tones:

Interval between medium level alarm tones:

Interval between low level alarm tones:

1 s

5 s

20 s

WARNING

When the alarm sound is switched off, the monitor will give no audible alarm tones even if a new alarm occurs. Therefore the user should be very careful about whether to switch off the alarm sound or not.

Do not rely exclusively on the audible alarm system for monitoring. Adjustment of alarm volume to a low level may result in a hazard to the patient. Always keep the patient under close surveillance.

7.4.4

Changing the Alarm Tone Pattern

To change the alarm tone pattern:

1.

Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2.

Select [

Alarm Setup >>

] to enter the [

Alarm Setup

] menu.

3.

Select [

Alarm Sound

] and then select [

ISO

], [

Mode 1

] or [

Mode 2

].

NOTE

User or factory default configurations have no impact on the setup of alarm tone pattern. The alarm tone pattern remains unchanged after the monitor restarts.

Passport 12m/Passport 17m Operator's Manual 7-5

7.4.5

Setting the Reminder Tones

When the alarm volume is set to zero, or the alarm is reset or turned off, the monitor issues a periodic reminder tone.

To set the reminder tone:

1.

Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2.

Select [

Alarm Setup >>

] to enter the [

Alarm Setup

] menu.

3.

Set the [

Reminder Tones

] to [

On

], [

Off

] or [

Re-alarm

]. When [

Re-alarm

] is selected, the acknowledged physiological alarms and technical alarms marked with “√” will be re-generated after the [

Reminder Interval

] if the alarm condition persists.

To set the interval between reminder tones, select [Reminder Interval] and toggle between [

1min

], [

2min

] and [

3min

].

In addition, you can set the volume of alarm reminder tones. To set the volume of alarm reminder tones, select [

Main

Menu

]

[

Alarm Setup >>

]

[

Others

] or the [

Alarm Setup

] QuickKey

[

Others

]. Then, select [

Reminder Vol

] and then select [

High

], [

Medium

] or [

Low

].

7.5

Understanding the Alarm Setup Menu

Select the [

Alarm Setup

] QuickKey or [

Main Menu

]

[

Alarm Setup >>

] to enter the [

Alarm Setup

], where you can:

Set alarm properties for all parameters.

Change ST alarm settings.

Change arrhythmia alarm settings.

 Set the threshold for some arrhythmia alarms.

 Change other settings.

The [

Alarm Setup

] menu as shown as below:

7-6

Refer to chapter

8 Monitoring ECG

for how to change ST alarm settings, how to change arrhythmia alarm settings and

how to set the threshold for some arrhythmia alarms.

Passport 12m/Passport 17m Operator's Manual

7.5.1

Setting Alarm Properties for All Parameters

In the [

Main Menu

], select [

Alarm Setup >>

]

[

Parameters

]. You can review and set alarm limits, alarm switches, alarm level and alarm recordings for all parameters.

When a measurement alarm occurs, automatic recording of all the measurement numerics and related waveforms is possible when the measurement’s [

On/Off

] and [

Record

] are set on.

WARNING

Make sure that the alarm limit settings are appropriate for your patient before monitoring.

Setting alarm limits to extreme values may cause the alarm system to become ineffective. For example, high oxygen levels may predispose a premature infant to retrolental fibroplasia. If this is a consideration, do NOT set the high alarm limit to 100%, which is equivalent to switching the alarm off.

When monitoring patients that are not continuously attended by a clinical operator, properly configure the alarm system and adjust alarm settings as per the patient's condition.

7.5.2

Adjusting Alarm Limits Automatically

The monitor can automatically adjust alarm limits according to the measured vital signs. When auto limits are selected, the monitor calculates safe auto limits based on the latest measured values.

To get accurate auto alarm limits, you need to collect a set of measured vital signs as a baseline. Then, in the [

Main

Menu

], select [

Alarm Setup >>

]

[

Parameters

]

[

Auto Limits

]

[

Ok

]. The monitor will create new alarm limits based on the measured values.

Before applying these automatically created alarm limits, confirm if they are appropriate for your patient in the [

Alarm

Setup

] menu. If not, you can adjust them manually. These alarm limits will remain unchanged until you select auto limits again or adjust them manually.

The monitor calculates the auto limits based on the following guidelines.

Low alarm limit High alarm limit

Module Parameter

ECG HR/PR

Adult/ pediatric

Neonate

Adult/ pediatric

Neonate

Auto alarm limits range

(HR × 0.8) or 40 greater)

(HR – 30) or 90 bpm (whichever is bpm (whichever is greater)

(HR × 1.25) or 240 bpm (whichever is bpm (whichever is

Adult/pediatric: 35 to 240

Neonate: 55 to 225 smaller)

(HR + 40) or 200 smaller)

Resp RR

(RR × 0.5) or 6 rpm

(whichever is greater)

(RR – 10) or 30 rpm (whichever is greater)

(RR × 1.5) or 30 smaller)

(RR + 25) or 85 rpm (whichever is rpm (whichever is smaller)

Adult/pediatric: 6 to 55

Neonate: 10 to 90

SpO

2

SpO

2

Same as the default alarm limit

Same as the default alarm limit

Same as the Same as the default alarm limit default alarm limit

Same as the measurement range

Passport 12m/Passport 17m Operator's Manual 7-7

Module Parameter

NIBP

NIBP-S

NIBP-D

NIBP-M

T1

Low alarm limit

Adult/ pediatric

(SYS × 0.68 + 10) mmHg

High alarm limit

Neonate

Adult/ pediatric

(SYS – 15) or 45 mmHg

(whichever is greater)

(SYS × 0.86 + 38) mmHg

(Dia × 0.68 + 6) mmHg

(Dia – 15) or 20 mmHg

(whichever is greater)

(Mean – 15) or

(Mean × 0.68 + 8) mmHg

35 mmHg

(whichever is greater)

(T1 – 0.5) °C (T1 – 0.5) °C

(Dia × 0.86 + 32) mmHg

(Mean × 0.86 +

35) mmHg

(T1 + 0.5) °C

Neonate

Auto alarm limits range

(SYS + 15) or 105 mmHg

(whichever is smaller)

Adult: 45 to 270

Pediatric: 45 to 185

Neonate: 40 to 115

(Dia + 15) or 80 mmHg

(whichever is

Adult: 25 to 210

Pediatric: 25 to 150

Neonate: 20 to 90 smaller)

(Mean + 15) or 95

Adult: 30 to 230 mmHg

Pediatric: 30 to 165

(whichever is

Neonate: 25 to 105 smaller)

(T1 + 0.5) °C 1 to 49 °C

T2 (T2 – 0.5) °C (T2 – 0.5) °C (T2 + 0.5) °C (T2 + 0.5) °C 1 to 49 °C

Temp

TD

IBP

IBP-S

Ao/

UAP/

BAP/

FAP/

LV/

P1-P4

(Arterial

IBP-D pressure)

IBP-M

Same as the default alarm limit

Same as the default alarm limit

(SYS × 0.68

mmHg

(Dia × 0.68

6)mmHg

(Mean × 0.68 +

8)mmHg

SYS × 0.75

Dia × 0.75

Mean × 0.75

10)

Same as the

(SYS – 15) or 45 mmHg

(whichever is greater)

(Dia – 15) or 20 mmHg

(whichever is greater)

(Mean – 15) or

35 mmHg

(whichever is greater)

(SYS × 0.86

38) mmHg

(Dia × 0.86

32)mmHg

(Mean × 0.86 +

35)mmHg

SYS × 0.75

Dia × 0.75

Mean × 0.75

SYS × 1.25

Dia × 1.25

Mean × 1.25

Same as the default alarm limit default alarm limit range

(SYS + 15) or 105 mmHg

(whichever is smaller)

(Dia + 15) or 80 mmHg

(whichever is smaller)

SYS × 1.25

Dia × 1.25

Mean × 1.25

Same as the measurement

Adult: 45 to 270

Pediatric: 45 to 185

Neonate: 35 to 115

Adult: 25 to 225

Pediatric: 25 to 150

Neonate: 20 to 90

(Mean + 15) or 95 mmHg

Adult: 30 to 245

Pediatric: 30 to 180

(whichever is smaller)

Neonate: 25 to 105

3 to 120 mmHg PA

IBP-S

IBP-D

IBP-M

IBP:

CVP/

ICP/

LAP/

RAP/

UVP/

P1-P4

(Venous pressure)

IBP-M Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25 3 to 40 mmHg

7-8 Passport 12m/Passport 17m Operator's Manual

Low alarm limit

Module Parameter

Adult/ pediatric

Neonate

0 to 32 mmHg:

0 to 32 mmHg: remains the same remains the same

32 to 35 mmHg:

29 mmHg

High alarm limit

Adult/ pediatric

Neonate

0 to 32 mmHg: 0 to 32 mmHg: remains the same remains the same

32 to 35 mmHg: 32 to 35 mmHg:

29 mmHg 41 mmHg

32 to 35 mmHg:

41 mmHg

Auto alarm limits range

CO

2

AG

C.O.

RM

EtCO

2

35 to 45 mmHg:

(etCO

2

-6) mmHg

35 to 45 mmHg: 35 to 45 mmHg:

(etCO

2

-6) mmHg (etCO

2

+6) mmHg

35 to 45 mmHg:

(etCO

2

+6) mmHg

Same as the measurement range

45 to 48 mmHg:39

45 to 48 mmHg mmHg:39 mmHg

>48 mmHg: remains the same

>48 mmHg: remains the same

45 to 48 mmHg:51 mmHg

>48 mmHg: remains the same

45 to 48 mmHg:51 mmHg

>48 mmHg: remains the same

FiCO

2

N/A N/A

Same as the Same as the Same as the measurement default alarm limit default alarm limit range awRR awRR × 0.5 or 6

(awRR – 10) or rpm (whichever is

30 rpm greater)

(whichever is greater)

EtCO

2

(AG)

FiCO

2

(AG)

Same as CO

2 module awRR awRR × 1.5 or 30 rpm (whichever is rpm (whichever is smaller)

(awRR+25) or 85 smaller)

(awRR – 10)

(awRR × 0.5) or 6 rpm (whichever is or 30 rpm

(whichever greater) is greater)

(awRR × 1.5) or 30

(awRR+25) or

85 rpm rpm (whichever is

(whichever is smaller) smaller)

Adult/pediatric: 6 to 55

Neonate: 10 to 90

Adult/pediatric: 6 to 55

Neonate: 10 to 90

FiAA/

EtAA

FiO

2

/

EtCO

2

FiN

2

O/

EtN

2

O

BT

RR(RM)

PEEP

Same as the default alarm limit

Same as the default alarm limit

Same as the default alarm limit

Same as the default alarm limit

Same as the

Same as the default alarm limit default alarm limit

Same as the

Same as the default alarm limit default alarm limit

Same as the

Same as the default alarm limit default alarm limit

Same as the

Same as the default alarm limit default alarm limit

Adult:

(BT – 1) °C

N/A

Adult:

(BT – 1) °C

N/A

(awRR × 0.5) or 6 rpm (whichever is N/A greater)

(PEEP – 5) cmH

2

O

N/A

(awRR × 1.5) or 30 rpm (whichever is N/A smaller)

(PEEP

5) cm

H

2

O

N/A

Same as the measurement range

Same as the measurement range

Same as the measurement range

Same as the measurement range

Adult/pediatric: 6 to 55

Neonate: 10 to 90

Same as the measurement range

Passport 12m/Passport 17m Operator's Manual 7-9

Low alarm limit

Module Parameter

Adult/ pediatric

PIP

MVe

Neonate

(PIP – 10) cmH

2

O

N/A

(MVe – 2) L/min N/A

BIS BIS N/A

CCO

CCO/

CCI, EDV/

EDVI, SVR/

N/A

SVRI,

SV/SVI,

RVEF

High alarm limit

Adult/ pediatric

Neonate

(PIP

10) cmH

2

O

N/A

(MVe

2) L/min N/A

Auto alarm limits range

Same as the measurement range

Same as the measurement range

SvO

2

SvO

2

ScvO

2

(SvO

2

– 5)%

(ScvO

2

– 5)%

N/A

N/A

(SvO

2

+ 5)%

(ScvO

2

+ 5)%

N/A

N/A

Same as the measurement range

Same as the measurement range

7.5.3

Setting Alarm Delay Time

You can set the alarm delay time for alarms of continuously measured parameters. If the alarm condition is resolved within the delay time, the monitor will not sound the alarm.

To set the alarm delay time:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]. Enter the required password and then select

[

Ok

].

2. Select [

Alarm Setup >>

]

[

Alarm Delay

].

Alarm delay is not applied to the following physiological alarms:

Apnea

ST alarms

Arrhythmia alarms

ECG Weak Signal

Resp Artifact

SpO

2

Desat

No Pulse

Nellcor SpO

2 over alarm limits

FiO

2

Shortage

Measurements of noncontinuous parameters over alarm limits

HR over alarm limits

 Anesthetic Mixture's MAC > 3

7-10 Passport 12m/Passport 17m Operator's Manual

You can set [

Apnea Delay

] and [

ST Alarm Delay

] separately.

To set the [

Apnea Delay

], select [

Main Menu

]

[

Alarm Setup >>

]

[

Others

].

To set the [

ST Alarm Delay

]:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]. Enter the required password and then select

[

Ok

].

2. Select [

Alarm Setup >>

]

[

Alarm Delay

].

7.5.4

Setting SpO2 Technical Alarm Delay

You can set the [

Tech. Alarm Delay

] in the [

Others

] tab of the [

Alarm Setup

] menu. The options are [

Off

], [

5 s

], [

10 s

] and [

15 s

]. The delay is effective to the following technical alarms: SpO

2

Sensor Off, SpO

2

Too Much Light, SpO

2

Low

Signal and SpO

2

Interference.

7.5.5

Setting Recording Length

You can change the length of the recorded waveforms. In the [

Others

] window of the [

Alarm Setup

] menu, select

[

Recording Length

] and then select [

8 s

], [

16 s

] or [

32 s

]:

 [

8 s

]: 4 seconds respectively before and after the alarm or manual event trigger moment.

 [

16 s

]: 8 seconds respectively before and after the alarm or manual event trigger moment.

 [

32 s

]: 16 seconds respectively before and after the alarm or manual event trigger moment.

7.5.6

Entering CPB Mode (Cardiopulmonary Bypass Mode)

When performing CPB, you can put the monitor in CPB mode in order to reduce unnecessary alarms. The CPB mode is activated only if you set the department to [

OR

]. To set the department to [

OR

]:

1. Press the [

Menu

] hardkey on the monitor’s front panel to enter [

Main Menu

].

2. Select [

Maintenance >>

]

[

Manage Configuration >>

]. Enter the required password and then select [

Ok

].

3. Select [

Change Department >>

]

[

OR

].

In the CPB mode, all the physiological alarms, technical alarms and prompt alarms are switched off except for BIS, tcGas, and NMT related alarms. In CPB mode, [

CPB Mode

] is displayed in the physiological alarm area with a red background color.

To enter CPB mode:

1. Select the [

CPB Mode

] Quickkey or select [

Enter CPB Mode

] in the [

Others

] window of the [

Alarm Setup

] menu.

2. Then select [

Ok

] in the pop-up dialog box.

Passport 12m/Passport 17m Operator's Manual 7-11

7.6

Pausing Alarms

You can temporarily disable alarm indicators by pressing the [

Alarm Pause

] hardkey on the monitor’s front panel. When alarms are paused:

For physiological alarms, no alarm indication is shown. New physiological alarms will not be presented.

The remaining alarm pause time is displayed in the physiological alarm area.

For technical alarms, alarm sounds are paused, but alarm lamps and alarm messages remain presented.

The alarms pause symbol is displayed in the sound symbol area. If a new technical alarm is triggered in the alarm paused period, the alarm message will be displayed.

If the time interval of the monitor’s last shutdown from this start-up is greater than 2 minutes, the monitor enters into the alarm paused status as soon as it is turned on. The alarm pause time is fixed to be 2 minutes.

When the alarm pause time expires, the alarm paused status is automatically cancelled and the alarm tone will sound.

You can also cancel the alarm paused status by pressing the [

Alarm Pause

] hardkey.

You can set the alarm pause time to [

1 min

], [

2 min

], [

3 min

], [

5 min

], [

10 min

], [

15 min

], or [

Permanent

]. The default alarm pause time is 2 minutes.

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2. Select [

Alarm Setup >>

]

[

Alarm Pause Time

] and then select the appropriate setting from the pop-up list.

7.7

Switching Off All Alarms

If [

Alarm Pause Time

] is set to [

Permanent

], the monitor will enter into the alarm off status after the [

Alarm Pause

] hardkey is pressed. During the alarm off status:

For physiological alarms, no alarm lamps flash and no alarms are sounded.

For physiological alarms, no numerics or alarm limits flash.

No physiological alarm messages are shown.

[

Alarm Off

] is displayed in the physiological alarm area with red background.

For technical alarms, no alarms are sounded.

The alarm off symbol is displayed in the alarm symbol area.

You can cancel the alarm off status by pressing the [

Alarm Pause

] hardkey.

WARNING

Pausing or switching off alarms may result in a hazard to the patient.

7-12 Passport 12m/Passport 17m Operator's Manual

7.8

Resetting Alarms

By selecting the QuickKey, you can reset the alarm system to acknowledge the on-going alarms and enable the alarm system to respond to a subsequent alarm condition.

For physiological alarms, except the NIBP-related alarms, when the alarm system is reset:

The alarm sound is silenced.

A

appears before the alarm message, indicating that the alarm is acknowledged.

The icon appears in the alarm symbol area.

 The parameter numeric and alarm limits still flash.

The indication of alarm lamp for the physiological alarm depends on the alarm light setting.

When [

Alarm Light on Alarm Reset

] is set to [

On

], the alarm lamp remains flashing.

When [

Alarm Light on Alarm Reset

] is set to [

Off

], the alarm lamp stops flashing.

To set [

Alarm Light on Alarm Reset

]:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2. Select [

Alarm Setup >>

] to enter the [

Alarm Setup

] menu.

3. Select [

Alarm Light on Alarm Reset

], and then select [

On

] or [

Off

].

NOTE

The default setting for [Alarm Light on Alarm Reset] is [On].

Technical alarms give different alarm indicators when the alarm system is reset:

 For some technical alarms, including the NIBP-related alarms, a

appears before the alarm message and appears in the alarm symbol area, indicating that the alarm is acknowledged. The indication of the alarm lamp depends on the alarm light setting.

Some technical alarms are changed to the prompt messages.

Some technical alarms are cleared. The monitor gives no alarm indications.

For details about the indications of technical alarms when the alarm system is reset, refer to appendix

D.2

Technical

Alarm Messages

.

Passport 12m/Passport 17m Operator's Manual 7-13

7.9

Latching Alarms

The latching setting for physiological alarms defines how alarm indicators behave if you do not reset the alarms.

 If you do not “latch” the physiological alarms, their alarm indications disappear when the alarm condition ends.

If you “latch” physiological alarms, all visual and audible alarm indications remain until you reset the alarms. For latched alarms the time when the alarm is last triggered is displayed beside the alarm message.

You can separately latch the visual indications or simultaneously latch the visual and the audible indications.

 When the visual indications are latched, the visual indications, including alarm lamp, alarm message and its background remains when the alarm condition ends.

When the audible indications are latched, the monitor issues alarm sounds when the alarm condition ends.

To latch a physiological alarm:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2. Select [

Alarm Setup

>>]

[

Latching Alarms>>

].

3. Select how you want to latch the alarms.

7-14

The rules for latching the alarms are:

You can separately select [

Latching Visual Signal

].

Selecting [

Latching Audible Signal

] simultaneously latches the visual signal.

Selecting alarms of lower priority simultaneously latches the alarms of higher priority.

The lethal alarms are latched by default. High, medium, and low priority alarms are unlatched by default.

NOTE

Changing of alarm priority may affect the latching status of corresponding alarms. Please determine if you need to adjust the latching status for the specific alarm when you have changed its alarm priority.

When the alarm system is reset, the latched physiological alarms are cleared.

You can separately configure the latching of the Vtac alarm. To do so, follow this procedure:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

Passport 12m/Passport 17m Operator's Manual

2. Select [

Alarm Setup

>>]

[

Latching Alarms>>

].

3. Set [

Vtac Latching Lock

] to [

Disable

].

4. Select [

Alarm Setup

>>] quickkey to enter [

Alarm Setup

] menu.

5. Select [

Others

] tab.

6. Set [

Vtac Latching

] as desired.

NOTE

If the visual signal of lethal alarms is not latched, [

Vtac Latching Lock

] is not configurable and is automatically set to [Enable].

7.10

Testing Alarms

When the monitor starts up, a self-test is performed. In this case the alarm lamp is lit in yellow and red respectively, and the system gives a beep. This indicates that the visible and audible alarm indicators are functioning correctly.

For further testing of individual measurement alarms, perform the measurement on yourself (for example, SpO

2

or CO

2

) or use a simulator. Then adjust alarm limits and check that appropriate alarm behavior is observed.

7.11

Using Care Group Alarms

7.11.1

Care Group Auto Alarms

When a Care Group is set up on your monitor, a flashing symbol will appear beside the QuickKeys area if any monitor in your Care Group, which is not currently viewed by your monitor, is alarming. The alarm symbol is shown as below.

When a monitor in the Care Group is disconnected, the flashing symbol is shown as below.

The department and bed label of the alarming monitor appear on the symbols. You can enter the View Other Patient window by pressing this symbol.

7.11.2

Setting Care Group Alert Tone

When a monitor in the Care Group issues an alarm, your monitor prompts you by sounding an alert tone. To set the alert tone:

1. In the [

Main Menu

], s]elect [

Screen Setup >>

]

[

Screen Layout >>

]

[

Choose Screen

].

2. In the [

Choose Screen

] window, select [

View Others Screen

].

3. In the View Other Patient window, select [

Setup>>

], and set [

Alert Tone

] to [

Repeat

], [

Once

] or [

Off

].

Passport 12m/Passport 17m Operator's Manual 7-15

7.11.3

Resetting Care Group Alarms

You can reset the alarm of the currently viewed bed in the View Other Patient window. This function can be set in the

[

Alarm Setup

] menu from the [

User Maintenance

] menu only.

To enable this function:

1. Select [

Main Menu

]

[

Maintenance>>

]

[

User Maintenance>>

]

→ enter the required password

[

Alarm

Setup>>

].

2. In the [

Alarm Setup

] menu, set [

Reset Other Bed’s Alarms

] to [

On

].

When your monitor is viewed by other monitors and [

Alarm Reset By Other Bed

] is switched on, alarms on your monitor can be reset by other monitors.

WARNING

Resetting care group alarms may cause a potential hazard.

7-16 Passport 12m/Passport 17m Operator's Manual

8

Monitoring ECG

8.1

Introduction

The electrocardiogram (ECG) measures the electrical activity of the heart and displays it on the monitor as a waveform and a numeric. This monitor measures ECG using the MPM module or the T1, or the N1. ECG monitoring provides two algorithms:

 Advanced ECG algorithm

The Advanced ECG algorithm enables 3-, 5- and 12-lead ECG monitoring, ST-segment analysis, and extended arrhythmia analysis.

Glasgow ECG algorithm

Glasgow algorithm provides resting 12-lead ECG analysis.

You can select algorithms at the time of purchase or as upgrades post sale:

 The MPM module or T1 incorporating the Advanced ECG algorithm is labeled with an ST/ARR label.

The N1 incorporating the ST feature is labeled with an ST label.

 The MPM module, the T1, or the N1 incorporating the Glasgow algorithm is labeled with the Glasgow logo.

8.2

Safety

WARNING

Use only ECG electrodes and cables specified by the manufacturer.

Make sure the conductive parts of electrodes and associated connectors for applied parts, including the neutral electrode, should not contact any other conductive parts including earth.

Periodically inspect the electrode application site to ensure skin quality. If the skin quality changes, replace the electrodes or change the application site.

Use defibrillation-proof ECG cables during defibrillation.

Do not touch the patient, or table, or instruments during defibrillation.

The neutral electrode of the HF surgical unit shall properly contact the patient. Otherwise, burns may result.

To reduce the hazard of burns during high-frequency surgical procedure, ensure that the monitor’s cables and transducers never come into contact with the electrosurgery unit (ESU).

The equipment is not intended for direct cardiac application.

CAUTION

Interference from a non-grounded instrument near the patient and electrosurgery interference can cause problems with the waveform.

Passport 12m/Passport 17m Operator's Manual 8-1

NOTE

After defibrillation, the screen display recovers within 10 seconds if the correct electrodes are used and applied in accordance with the manufacturer’s instructions for use.

8.3

Preparing to Monitor ECG

8.3.1

Preparing the Patient and Placing the Electrodes

To prepare the patient and place the electrodes:

1.

Prepare the patient’s skin. Proper skin preparation is necessary for good signal quality at the electrode site, as the skin is a poor conductor of electricity. To properly prepare the skin, choose flat areas and then:

Shave hair from skin at chosen sites.

Gently rub skin surface at sites to remove dead skin cells.

Thoroughly cleanse the site with a mild soap and water solution. We do not recommend using ether or pure alcohol, because this dries the skin and increases the resistance.

Dry the skin completely before applying the electrodes.

2.

Attach the clips or snaps to the electrodes before placing them.

3.

Place the electrodes on the patient.

4. Attach the electrode cable to the patient cable and then plug the patient cable into the ECG connector on the MPM, the T1, or the N1.

8.3.2

Choosing AHA or IEC Lead Placement

To choose AHA or IEC lead placement:

1.

Select the ECG parameter window or waveform area to enter the [

ECG Setup

] menu.

2.

Select [

Others

]

[

Lead Set

] and then select [

3-lead

], [

5-lead

], [

12-lead

] or [

Auto

] according to the applied electrodes.

3.

Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

4.

Select [

Others >>

]

[

ECG Standard

] and then select [

AHA

] or [

IEC

] according to the hospital standard.

8-2 Passport 12m/Passport 17m Operator's Manual

8.3.3

ECG Lead Placements

The electrode placement illustrations in this chapter adopt the AHA standard.

3-Leadwire Electrode Placement

Following is an electrode configuration when using 3 leadwires:

RA placement: directly below the clavicle and near the right shoulder.

LA placement: directly below the clavicle and near the left shoulder.

LL placement: on the left lower abdomen.

5-Leadwire Electrode Placement

Following is an electrode configuration when using 5 leadwires:

RA placement: directly below the clavicle and near the right shoulder.

LA placement: directly below the clavicle and near the left shoulder.

RL placement: on the right lower abdomen.

LL placement: on the left lower abdomen.

V placement: on the chest.

The chest (V) electrode can be placed on one of the following positions:

V1 placement: on the fourth intercostal space at the right sternal border.

 V2 placement: on the fourth intercostal space at the left sternal border.

V3 placement: midway between the V2 and V4 electrode positions.

V4 placement: on the fifth intercostal space at the left midclavicular line.

 V5 placement: on the left anterior axillary line, horizontal with the V4 electrode position.

V6 placement: on the left midaxillary line, horizontal with the V4 electrode position.

V3R-V6R placement: on the right side of the chest in positions corresponding to those on the left.

 VE placement: over the xiphoid process.

V7 placement: on posterior chest at the left posterior axillary line in the fifth intercostal space.

V7R placement: on posterior chest at the right posterior axillary line in the fifth intercostal space.

Passport 12m/Passport 17m Operator's Manual 8-3

12-Leadwire Electrode Placement

12-lead ECG uses 10 electrodes, which are placed on the patient’s four limbs and chest. The limb electrodes should be placed on the soft skin and the chest electrodes placed according to the physician’s preference.

Lead Placement for Surgical Patients

The surgical site should be taken into consideration when placing electrodes on a surgical patient. e.g. for open-chest surgery, the chest electrodes can be placed on the lateral chest or back. To reduce artifact and interference from electrosurgical units, you can place the limb electrodes close to the shoulders and lower abdomen and the chest electrodes on the left side of the mid-chest.

Do not place the electrodes on the upper arm. Otherwise, the ECG waveform will be very small.

WARNING

When using electrosurgical units (ESU), patient leads should be placed in a position that is equal distance from the electrosurgery electrotome and the grounding plate to avoid burns to the patient. Never entangle the ESU cable and the ECG cable together.

When using electrosurgical units (ESU), never place ECG electrodes near to the grounding plate of the ESU, as this can cause a lot of interference on the ECG signal.

8.3.4

Checking Paced Status

It is important to set the paced status correctly when you start monitoring ECG. The paced symbol is displayed in the ECG waveform area when the [

Paced

] status is set to [

Yes

]. The pace pulse markers “|” are shown on the ECG wave when the patient has a paced signal. If [

Paced

] is set to [

No

] or the patient’s paced status is not selected, the symbol

8-4

will be shown in the ECG waveform area.

To change the paced status, you can select either:

 the patient information area, or

[

Main Menu

]→[

Patient Setup

]→[

Patient Demographics

], or,

 the ECG parameter window or waveform area

→[

Others >>

], and then, select [

Paced

] from the pop-up menu and then select [

Yes

] or [

No

].

If you do not set the paced status, the monitor issues a prompt tone when pace pulse is detected. At the same time, the paced symbol flashes and the message [

Please confirm the pace of patient

] appears in the ECG waveform area. Then, please check and set the paced status of the patient.

Passport 12m/Passport 17m Operator's Manual

WARNING

For paced patients, you must set [Paced] to [Yes]. If it is incorrectly set to [No], the monitor could mistake a pace pulse for a QRS and fail to alarm when the ECG signal is too weak.

On ventricular paced patients, episodes of Ventricular Tachycardia may not always be detected.

False low heart rate indicators or false Asystole calls may result with certain pacemakers because of pacemaker artifact such as electrical overshoot of the pacemaker overlapping the true QRS complexes.

Do not rely entirely upon the system’s automated arrhythmia detection algorithm. Keep pacemaker patients under close surveillance.

For non-paced patients, you must set [Paced] to [No].

The auto pacer recognition function is not applicable to pediatric and neonatal patients.

8.4

Understanding the ECG Display

Your display may be configured to look slightly different.

1 2 3

1. Lead label of the displayed wave

2. ECG gain

3. ECG filter label

4

4. Notch filter status

Additionally, when a paced signal has been detected, the pace pulse marks “|” are shown on the ECG wave if the [

Paced

] setting has been configured to [

Yes

].

3

1

2

1. Current heart rate alarm limits

2. Current heart rate

3. Heart beat symbol

Passport 12m/Passport 17m Operator's Manual 8-5

8.5

Changing ECG Settings

8.5.1

Accessing ECG Menus

By selecting the ECG parameter window or waveform area, you can access the [

ECG Setup

] menu.

8.5.2

Choosing the Alarm Source

In most cases the HR and PR numerics are identical. In order to avoid simultaneous alarms on HR and PR, the monitor uses either HR or PR as its active alarm source. To change the alarm source, select [

Alm Source

] in the [

ECG Setup

] menu and then select either:

[

HR

]: if you want the HR to be the alarm source for HR/PR.

[

PR

]: if you want the PR to be the alarm source for HR/PR.

[

Auto

]: If the [

Alm Source

] is set to [

Auto

], the monitor will use the heart rate from the ECG measurements as the alarm source whenever a valid heart rate is available. If the heart rate becomes unavailable, for example the ECG module is turned off or becomes disconnected, the monitor will automatically switch to PR as the alarm source.

8.5.3

Changing ECG Wave Settings

In the [

ECG Setup

] menu:

You can select [

ECG

], [

ECG1

], or [

ECG2

] to select a lead to view. The ECG lead(s) selected is used for analysis (beat detection, arrhythmia classification, and V-Fib detection). The waveform of selected lead should have the following characteristics:

The QRS should be either completely above or below the baseline and it should not be biphasic.

The QRS should be tall and narrow.

The P-waves and T-waves should be less than 0.2mV.

If the wave is too small or clipped, you can change its size by selecting an appropriate [

Gain

] setting. If you select

[

Auto

] from [

Gain

], the monitor will automatically adjust the size of the ECG waves. In normal screen, only the selected ECG wave’s size is adjusted. In other screens, all ECG waves’ size is adjusted simultaneously.

You can change the wave sweep speed by selecting [

Sweep

] and then selecting the appropriate setting.

NOTE

Changing the display gain on the monitor does not affect the amplitude of signal that is used by the algorithm for beat detection.

8-6 Passport 12m/Passport 17m Operator's Manual

8.5.4

Changing the ECG Filter Settings

The ECG filter setting defines how ECG waves are smoothed. To change the filter setting, select [

Filter

] from [

ECG Setup

] and then select the appropriate setting.

[

Monitor

]: Use under normal measurement conditions.

[

Diagnostic

]: Use when diagnostic quality is required. The unfiltered ECG wave is displayed so that changes such as

R-wave notching or discrete elevation or depression of the ST segment are visible.

[

Surgery

]: Use when the signal is distorted by high frequency or low frequency interference. High frequency interference usually results in large amplitude spikes making the ECG signal look irregular. Low frequency interference usually leads to wandering or rough baseline. In the operating room, the surgery filter reduces artifacts and interference from electrosurgical units. Under normal measurement conditions, selecting [

Surgery

] may suppress the QRS complexes too much and then interfere with ECG analysis.

[

ST

]: Recommended when ST monitoring is used. The unfiltered ECG wave is displayed so that changes such as discrete elevation or depression of the ST segment are visible.

CAUTION

The [Diagnostic] filter is recommended when monitoring a patient in an environment with slight interference only.

8.5.5

Setting the ECG Lead Set

You can set the [

Lead Set

] by selecting [

ECG Setup

]

[

Others>>

]. You can set the [

Lead Set

] as [

Auto

] if the auto lead detection function is available.

8.5.6

Choosing an ECG Display Screen

When monitoring with a 5-lead or 12-lead set, you can select the [

Screens

] Quickkey and then in the [

Choose Screen

] window, choose the screen type as:

 [

Normal Screen

]: The ECG waveform area displays 2 ECG waveforms.

 [

ECG 7-Lead Full-Screen

]: The whole waveform area displays 7 ECG waveforms only.

 [

ECG 7-Lead Half-Screen

]: The upper half part of the whole waveform area displays 7 ECG waveforms.

 [

ECG 12-Lead Full-Screen

]: The whole waveform ares displays 12 ECG waveforms (for 12-lead set only).

When monitoring with a 3-lead set, the screen type can only be set to [

Normal

].

When the screen type is set to [

Normal Screen

] and [

Sweep Mode

] is set to [

Refresh

], cascaded ECG waveforms can be displayed.

To cascade ECG waveforms:

1. Select the [

Screens

] Quickkey

[

Screen Setup

].

2. Select [

ECG1 Casc

.] in the second row. A cascaded waveform is displayed in two waveform positions.

Passport 12m/Passport 17m Operator's Manual 8-7

8.5.7

Setting the Notch Filter

The notch filter removes the line frequency interference. Only when [

Filter

] is set to [

Diagnostic

], the [

Notch Filter

] is adjustable.

To set the notch filter:

1. Select the ECG parameter window or waveform area to enter its setup menu. Then select [

Others >>

].

2. Set [

Notch Filter

] to:

[

Strong

]: when waveform interference is strong (such as spikes).

[

Weak

]: when waveform interference is weak.

[

Off

]: to turn the notch filter off.

Set notch frequency according to the electric power frequency of your country. To set notch filter frequency:

1. When [

Notch Filter

] is turned on, select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2. Select [

Others >>

]

[

Notch Freq.

] and then select [

50 Hz

] or [

60 Hz

] according to the power line frequency.

8.5.8

Changing the Pacer Reject Settings

To change the pacer reject settings, select [

ECG Setup

]

[

Others>>

]

[

Pacer Reject

], and then select [

On

] or [

Off

].

When [

Paced

] is set to [

Yes

]:

When [

Pacer Reject

] is switched on, the pace pulses are not counted as extra QRS complexes.

The pace pulse marks “|” are shown on the ECG wave when pace pulses are detected.

When [

Paced

] is set to [

No

], the pace markers are not shown on the ECG wave, and the options of [

Pacer Reject

] are inactivated.

8.5.9

Enabling Smart Lead Off

When the smart lead off function is enabled and there is a “lead off” in the lead of the first ECG wave, if another lead is available, the monitor will automatically select the available lead. The system will re-calculate HR and analyze and detect arrhythmia. When the “lead off” condition is corrected, the leads are automatically switched back.

To switch on/off the smart lead off function, select [

Others >>

] from the [

ECG Setup

] menu; select [

Smart Lead Off

] and then select [

On

] or [

Off

] from the pop-up menu.

8.5.10

Setting the Alarm Level for ECG Lead Off Alarms

Select [

Alarm Setup >>

] from the [

User Maintenance

] menu. You can set [

ECGLeadOff Lev.

] from the pop-up menu.

8-8 Passport 12m/Passport 17m Operator's Manual

8.5.11

Adjusting QRS Volume

QRS sounds are produced based on the alarm source.

To adjust the QRS volume:

1. Select [

Others >>

] from the [

ECG Setup

] menu.

2. Select [

QRS Volume

] from the pop-up menu and select the appropriate setting.

When a valid SpO

2

measured value is available, the system adjusts the pitch tone of QRS sound based on the SpO

2

value.

8.5.12

About the Defibrillator Synchronization

A defibrillator synchronization pulse (100 ms, +5V) is output through the Defib. Sync connector every time the monitor detects an R-wave.

To use the defibrillator synchronization function, connect the monitor and the defibrillator with a synchronization cable.

WARNING

Improper use of a defibrillator may cause injury to the patient. The user should determine whether to perform defibrillation or not according to the patient’s condition.

8.5.13

Adjusting the Minimum QRS Detection Threshold (For Advanced ECG Algorithm

Only)

To avoid false asystole alarms when the R wave amplitude is low and missed asystole alarms during ventricular standstill

(tall P waves, but no QRS), a means to manually adjust the minimum QRS detection threshold is provided.

To adjust the QRS threshold,

1. In the [

ECG Setup

] menu, set [

Filter

] to [

Monitor

].

2. Select [Others >>]

[Minimum QRS Threshold >>] to enter the [Minimum QRS Threshold] menu.

3. Select the up or down arrow to adjust the QRS threshold. Selecting [

Default

] resets the QRS threshold to the default value (0.16 mV).

4. Select [

Confirm

] to make the changes effective.

CAUTION

The setting of QRS threshold can affect the sensitivity of arrhythmia, ST, QT/QTc detection, and heart rate calculation.

If QRS amplitude is low, the monitor might not be able to calculate heart rate and false asystole may occur.

Passport 12m/Passport 17m Operator's Manual 8-9

NOTE

The minimum QRS detection threshold can only be adjusted when the ECG filter is set to [Monitor].

8.6

About ST Monitoring

ST segment analysis calculates ST segment elevations and depressions for individual leads and then displays them as numerics in the ST1 and ST2 areas.

 A positive value indicates ST segment elevation; a negative value indicates ST segment depression.

Measurement unit of the ST segment deviation can be set to mV or mm. You can set the unit in the [

Unit Setup

] menu from the [

User Maintenance

] menu.

Measurement range of the ST segment deviation: -2.0 mV to +2.0 mV (-20.0 mm to +20.0 mm).

NOTE

The monitor gives the message “Cannot Analyze ST” when the monitor cannot generate a valid ST value. In this case, check the ECG signal quality and ST measurement points. If the message persists, check the

patient condition as per section

8.6.1Switching ST Monitoring On and Off

and decide whether to continue ST

monitoring.

The ST algorithm has been tested for accuracy of the ST segment data. The significance of the ST segment changes need to be determined by a clinician.

8.6.1

Switching ST Monitoring On and Off

To switch ST monitoring on or off:

1. In the [

ECG Setup

] menu, select [

ST Analysis >>

].

2. Next to the

ST Analysis

setting, select [

On

] or [

Off

].

Reliable ST monitoring cannot be ensured if:

 You are unable to get a lead that is not noisy.

 Arrhythmias such as atrial fib/flutter cause irregular baseline.

 The patient is continuously ventricularly paced.

 The patient has left bundle branch block.

In these cases, you may consider switching ST monitoring off.

8-10 Passport 12m/Passport 17m Operator's Manual

8.6.2

Understanding the ST Display

8.6.2.1

ST Numerics

This example shows ST numerics with 5-lead ECG. Your monitor screen may look slightly different from the illustration.

You can also change the ST numerics display.

To change the ST numerics display:

1. Enter the [

ST Analysis

] menu. Set

ST Analysis

to [

On

].

2. Set [

Display Leads

] to:

[

All

]: ST numercis of all ECG leads are displayed.

[

Selected

]: ST numercis of ECG1 and ECG2 leads set in the [

ECG Setup

] menu are displayed.

8.6.2.2

ST Segment

ST segment shows a QRS complex segment for each measured ST lead. The current ST segment is drawn in the same color as the ECG wave, usually green, superimposed over the stored reference segment, drawn in a different color. The information is updated once every ten seconds.

To display the ST segment on the normal screen:

1. Enter the [

ST Analysis

] menu. Set

ST Analysis

to [

On

].

2. Enter the [

Screen Setup

] window from the [

Screens

] menu. Set [

ST Segment

] to be displayed.

Passport 12m/Passport 17m Operator's Manual 8-11

Select the ST parameter window or ST segment area to enter the [

ST Analysis

] menu.

8.6.3

Saving the Current ST Segment as Reference

Select [

Save Ref.

] in the [

ST Analysis

] menu to save the current segment as a reference. Up to 20 reference segment groups can be saved.

NOTE

If the memory is full and you do not delete a group before saving a new one, the earliest saved group is deleted automatically.

8.6.4

Changing the Reference Segment

Select the and arrow keys beside the [

Change Ref.

] to switch between different reference segment groups.

8.6.5

Deleting a Reference Segment

Select [

Delete Ref.

] in the [

ST Analysis

] menu and then select [

Ok

] in the pop-up to delete the current ST reference segment.

8.6.6

Recording the ST Segment

Select [

Record

] in the [

ST Analysis

] menu to record the current ST segment and reference segment.

8-12 Passport 12m/Passport 17m Operator's Manual

8.6.7

Changing the ST Alarm Limits

High and low ST alarm limits can be set individually for each ECG lead. Alarm limits can also be set separately for single-lead and multi-lead ST monitoring. You can select [

ST Alarm Setup >>

] from [

ST Analysis

] menu and then change ST alarm settings for each lead.

8.6.8

Setting the ST Alarm Delay Time

To set the ST alarm delay time:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]. Enter the required password and then select

[

OK

].

2. Select [

Alarm Setup >>

] → [

ST Alarm Delay

].

8.6.9

Adjusting ST Measurement Points

As shown in the figure below, the ST measured for each beat complex is the vertical difference between two measurement points with the R-wave peak as the baseline for the measurement.

R-wave peak

P J point

T

Q point

Isoelectric point

ST measurement point

Difference=ST value

The ISO and ST points need to be adjusted when you start monitoring and if the patient’s heart rate or ECG morphology changes significantly. Exceptional QRS complexes are not considered for ST-segment analysis.

WARNING

Always ensure that the ST measurement points are appropriate for your patient.

To adjust the ST measurement points:

1. Select the ECG numeric area or waveform area to enter the [

ECG Setup

] menu.

2

Select [

ST Analysis

>>].

3. Set

ST Analysis

to [

On

].

4. Select [

Adjust ST Point >>

].

Passport 12m/Passport 17m Operator's Manual 8-13

In the [

Adjust ST Point

] window, three vertical lines represent the ISO, J and ST point positions respectively. In this window, you can:

 Select [

Auto

] to automatically detect the ISO point and J point, and set ST point. When [

Auto

] is selected, the monitor takes a few seconds to finish automatic detection. During this time, the message [

Please wait..

.] is displayed in the [

Adjust ST Point

] window.

Or manually adjust the ISO point, J point, and ST point.

To manually adjust the ST measurement points, follow this procedure:

1. Use the left and right arrows beside [

ISO

] and [

J

] to adjust the position of ISO point and J point.

The ISO-point (isoelectric) position is given relative to the R-wave peak. Position the ISO-point in the middle of the flattest part of the baseline (between the P and Q waves).

The J-point position is given relative to the R-wave peak and helps locating the ST-point. Position the J-point at the end of the QRS complex and the beginning of the ST segment.

2. Set [

ST Point

]. The ST-point is positioned a fixed distance from the J-point. Move the J-point to position the

ST-point at the midpoint of the ST segment. Position the ST-point relative to the J-point at either [

J+60/80ms

], [

J+40ms

],

[

J+60ms

] or [

J+80ms

]. When [

J+60/80ms

] is selected, the ST-point will be positioned 80 ms (heart rate 120 bpm or less) or 60 ms (heart rate more than 120 bpm) from the J-point.

NOTE

Only Advanced ECG algorithm has the function of automatic J point detection.

The ISO point, J point, and ST point displayed in the [Adjust ST Point] window are positioned with reference to the R-wave peak.

8.7

QT/QTc Interval Monitoring (For Advanced ECG Algorithm Only)

The QT interval is defined as the time between the beginning of the Q-wave and the end of the T-wave. It measures the total duration of the depolarization (QRS duration) and repolarization (ST-T) phases of the ventricles. QT interval monitoring can assist in the detection of long QT syndrome.

The QT interval has an inverse relationship to heart rate. As heart rate increases, the QT interval shortens, while at lower heart rates QT interval gets longer. Several formulas are available to correct QT interval for heart rate. The heart rate corrected QT interval is abbreviated as QTc.

QT/QTc Interval Monitoring is intended for adult, pediatric, and neonate patients.

8-14

QT interval

Passport 12m/Passport 17m Operator's Manual

CAUTION

For QT interval monitoring to be effective, basic or enhanced arrhythmia monitoring should be turned on.

8.7.1

QT/QTc Monitoring Limitations

Some conditions may make it difficult to achieve reliable QT monitoring, for example:

 R-wave amplitudes are too low

The presence of frequent ventricular ectopic beats

 Unstable RR intervals

P-waves tending to encroach on the end of the previous T-wave at high heart rates

T-waves are very flat or not well defined The end of the T-wave is difficult to delineate because of the presence of

U-waves

QTc measurements are not stable

In the presence of noise, asystole, ventricular fibrillation, and ECG lead off

For these cases you should select a lead with good T-wave amplitude and no visible flutter activity, and without a predominant U-wave or P-wave.

Some conditions such as left or right bundle branch block or hypertrophy can lead to a widened QRS complex. If a long

QTc is observed you should verify it to ensure that it is not caused by QRS widening.

Because normal beats followed by ventricular beats are not included in the analysis, no QT measurement will be generated in the presence of a bigeminy rhythm.

If the heart rate is extremely high (over 150bpm for adults and over 180bpm for pediatrics and neonates), QT will not be measured. When the heart rate changes, it can take several minutes for the QT interval to stabilize. For reliable QTc calculation it is important to avoid the region where the heart rate is changing.

WARNING

QT/QTc measurements should always be verified by a qualified clinician

8.7.2

Enabling QT/QTc Monitoring

The QT monitoring function is disabled by default. Before you start QT monitoring, enable the QT function.

To enable QT/QTc monitoring:

1. In the [

ECG Setup

] menu, select [

QT Analysis>>

] to enter the [

QT Analysis

] menu.

2. Set [

QT Analysis

] to [

On

].

Passport 12m/Passport 17m Operator's Manual 8-15

8.7.3

Displaying QT/QTc Parameters and Waveform

To display QT numerics and waveform:

1. Select the [

Screens

] QuickKey or select [

Main Menu

]

[

Screen Setup

>>]

[

Screen Layout

>> to enter the

[

Screens

] window

2 Select the parameter area where you want to display the QT parameters and select [

QT

].

The following picture shows the QT numeric area. Your monitor screen may look slightly different:

3

4

1

1: Parameter label

2

2: QTc alarm limit (if QTc alarm is off, the alarm off symbol is displayed)

3: QTc value

5

6

4: ΔQTc value (the difference between the current and reference QTc values. If ΔQTc alarm is off, the alarm off symbol is displayed on the right.)

5: QT value

6: QT-HR value

NOTE

QTc values are calculated based on the QT-HR, not the ECG HR.

8.7.4

Accessing the QT View

QT View shows the current and reference QT parameter values and waveforms.

To access the QT View:

1. Select the QT parameter area or waveform area to enter the [

QT Analysis

] menu.

2. Select [

QT View>>

].

8-16 Passport 12m/Passport 17m Operator's Manual

The following picture shows the QT view.

 The current waveform is shown in the upper half in green.

 The reference waveform is shown below in yellow.

 The start of QRS complex and the end of the T wave are marked with vertical lines.

In some conditions, no QT measurement can be calculated. Then the cause of failed QT measurement is shown at the bottom of the QT numerics area. Additionally the message “Cannot Analyze QT” is shown in the technical alarm area.

In the QT View, you can:

Select the arrows besides [

View Leads

] to switch the displayed lead.

Select [

Save Ref.

] to save the current QT parameters and waveforms as reference. If no reference has been set for this patient within the first five minutes after getting valid QT values, the monitor will automatically set a reference. If you set a new reference, the previous reference will be discarded.

CAUTION

Updating QTc reference affects ΔQTc value and alarm.

8.7.5

Changing QT Settings

8.7.5.1

Setting QT Alarm Properties

To set QT alarm properties,

1. Select the [

Alarm Setup

] QuickKey, or select [

Alarm Setup>>

] from the

[

QT Analysis ]

menu.

2. Set QTc and ΔQTc alarm properties.

Passport 12m/Passport 17m Operator's Manual 8-17

8.7.5.2

Selecting the QT Analysis Lead

You can select one lead or all leads for QT calculation. To select the QT analysis lead:

1. Select the QT parameter area or waveform area to enter the [

QT Analysis

] menu.

2. Set [

Analysis Lead

].

When [

Analysis Lead

] is set to [

All

]:

For 12-lead ECG, leads I, II, III, V2 and V5 are displayed.

For 5-lead ECG, leads I, II, III, and V are displayed.

8.7.5.3

Changing QTc Formula

The monitor uses as a default the Hodges correction formula to correct the QT interval for heart rate. To change the QTc formula:

1. Select the QT parameter area or waveform area to enter the [

QT Analysis

] menu.

2. Set [

QTc Formula

].

Hodges:

QTc

=

QT

+ 1 .

75 ×

(

HeartRate

− 60

)

Bazett:

QTc

=

QT

×

HeartRate

60

1

2

Fridericia:

QTc

=

QT

×

HeartRate

60

1

3

Framingham:

QTc

=

QT

+ 154 ×

60

HeartRate

8.8

About Arrhythmia Monitoring

Arrhythmia analysis provides information about the patient's cardiac parameters like heart rate, PVC rate, and rhythm.

WARNING

Arrhythmia analysis program is intended to detect ventricular arrhymias and atrial fibrillation. It is not designed to detect all the atrial or supraventricular arrhythmias. It may incorrectly identify the presence or absence of an arrhythmia. Therefore, a physician must analyze the arrhythmia information with other clinical findings.

The arrhythmia detection is not intended for neonatal patients.

Heart-rate reading may be affected by cardiac arrhythmias. Do not rely entirely on rate meter alarms when monitoring patients with arrhythmias. Always keep these patients under close surveillance.

Atrial fibrillation (Afib) detection function is not intended for pediatric and neonatal patients.

8-18 Passport 12m/Passport 17m Operator's Manual

8.8.1

Understanding the Arrhythmia Events

Advanced ECG algorithm

Arrhythmia message

Asystole

Vfib/Vtac

Vtac

Vent. Brady

Extreme Tachy

Extreme Brady

PVCs

Pacer Not Pacing

Pacer Not Capture

PVC

Couplet

Run PVCs

Bigeminy

Trigeminy

R on T

Missed Beats

Brady

Tachy

Vent. Rhythm

Multif. PVC

Nonsus. Vtac

Pause

Irr. Rhythm

Afib

Description

No QRS detected within the set time threshold in absence of ventricular fibrillation or chaotic signal.

A fibrillatory wave for 6 consecutive seconds.

A dominant rhythm of adjacent Vs and a HR > the V-Tac HR limit.

The consecutive PVCs ≥ Vtac PVCs limit, and the HR ≥ the Vtac rate limit.

The consecutive PVCs ≥ the Vbrd threshold and the ventricular HR < the

Vbrd Rate threshold.

The heart rate is equal to or greater than the extreme tachycardia limit.

The heart rate is equal to or less than the extreme bradycardia limit.

PVCs/min exceeds high limit

No pace pulse detected for 1.75 x average R-to-R intervals following a

QRS complex (for paced patients only).

No QRS complex detected for 300 milliseconds following a pace pulse

(for paced patients only).

One PVC detected in normal heartbeats.

Paired PVCs detected in normal heartbeats.

More than 2 consecutive PVCs.

A dominant rhythm of N, V, N, V, N, V.

A dominant rhythm of N, N, V,N, N, V, N, N, V.

R on T detected in normal heartbeats.

No beat detected for 1.75 x average R-R interval for HR <120, or

No beat for 1 second with HR > 120 (for non-paced patients only), or

No beat detected for more than the set pause threshold.

The average heart rate is equal to or less than the bradycardia limit.

The average heart rate is equal to or greater than the tachycardia limit.

The consecutive PVCs ≥ the Vbrd PVCs limit, and HR is ≥ Vbrd Rate limit but < the Vtac Rate limit.

Multiform PVCs detected in Multif. PVC's Window (which is adjustable).

The consecutive PVCs < the Vtac PVCs limit but > 2, and HR ≥ the Vtac

Rate limit.

No QRS detected within the set time threshold of pause.

Consistently irregular rhythm.

Absence of P-waves and/or variable RR intervals between normal beats.

Category

Lethal arrhythmia

Nonlethal arrhythmia

Passport 12m/Passport 17m Operator's Manual 8-19

8.8.2

Changing Arrhythmia Alarm Settings

To change arrhythmia alarm settings, select the ECG parameter area or waveform area

[

ECG Setup

]

[

Arrh.

Analysis >>

]. In the pop-up menu, you can set the [

Alm Lev

] to [

High

], [

Med

], [

Low

] or [

Message

], or switch on lethal arrhythmia analysis alarms only or switch on/off all arrhythmia analysis alarms. In the [

Alarm Setup

] menu from the

[

User Maintenance

] menu, you can enable/disable turning off lethal arrhythmia analysis alarms.

WARNING

If you switch off all arrhythmia analysis alarms, the monitor cannot give any arrhythmia analysis alarm.

Always keep the patient under close surveillance.

NOTE

The priority of lethal arrythmia alarms is always high. It is unchangeable.

8.8.3

Changing Arrhythmia Threshold Settings

Select the ECG parameter window or waveform area

[

Arrh. Analysis >>

]

[

Arrh. Threshold

], and you can then change threshold settings for some arrhythmia alarms. When an arrhythmia violates its threshold, an alarm will be triggered. The asystole delay time relates to ECG relearning. When HR is less than 30 bpm, it is recommended to set the asystole delay time to 10 seconds.

Advanced ECG algorithm

Arrh. event

PVCs High

Asys. Delay

Tachy High

Brady Low

Extreme Tachy

Extreme Brady

Multif. PVC's Window

Vtac Rate

Vtac PVCs

Pause Time

Vbrd PVCs

Vbrd Rate

Range

1 to 100

3 to 10

60 to 300

15 to 120

120 to 300

15 to 60

3 to 31

100 to 200

3 to 99

1.5, 2.0,2.5

3 to 99

15 to 60

Default

10

5

Adult: 120

Pediatric: 160

Adult: 50

Pediatric: 75

Adult: 160

Pediatric: 180

Adult: 35

Pediatric: 50

6

2

15

130

5

40

5

5

5

1

/

1

5

1

5

Step

1

1

5

/min bpm

/min s

/min bpm

Unit

/min s bpm bpm bpm bpm

8-20 Passport 12m/Passport 17m Operator's Manual

8.8.4

Setting the Extended Arrhythmia (For Advanced ECG Algorithm Only)

The following arrhythmia events are defined as extended arrhythmia:

Extreme Tachy

Extreme Brady

 Vent. Brady

 Nonsus. Vtac

 Multif. PVC

 Irr. Rhythm

 Pause

 Afib

You can select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password

→ select

[

Alarm Setup >>

], and set [

Extended Arrh

.] to [

Enable

] or [

Disable

]. When [

Extended Arrh

.] is set to [

Disable

], the monitor does not analyze the extended arrhythmia events, and corresponding alarms are not given.

8.8.5

Reviewing Arrhythmia Events

Please refer to chapter

27 Review

.

8.9

ECG Relearning

8.9.1

Initiating an ECG Relearning Manually

During ECG monitoring, you may need to initiate an ECG relearning when the patient’s ECG template changes dramatically. A change in the ECG template could result in:

Incorrect arrhythmia alarms

Loss of ST measurement, and/or

Inaccurate heart rate

ECG relearning allows the monitor to learn the new ECG template to correct arrhythmia alarms and HR value, and restore

ST measurements. To initiate relearning manually, select the ECG parameter window or waveform area

[

Relearn

].

When the monitor is learning, the message [

ECG Learning

] is displayed in the technical alarm area.

WARNING

Initiate ECG relearning only during periods of normal rhythm and when the ECG signal is relatively noise-free. If ECG learning takes place during ventricular rhythm, the ectopics may be incorrectly learned as the normal QRS complex. This may result in missed detection of subsequent events of V-Tach and V-Fib.

Passport 12m/Passport 17m Operator's Manual 8-21

8.9.2

Automatic ECG Relearning

ECG relearning is initiated automatically whenever:

The ECG lead or lead label is changed.

The ECG lead is re-connected.

 A new patient is admitted.

 After ECG calibration is completed, and [

Stop Calibrating ECG

] is selected.

 A switch happens between the options of screen type during 5/12-lead ECG monitoring.

 The paced status of the patient is changed.

8.10

12-lead ECG Monitoring

To access the 12-lead ECG monitoring screen:

1. Refer to the section

8.3.3 ECG Lead Placements

to place the electrodes.

2. In the [

ECG Setup

] menu, select [

Others>>

] to enter the [

Other Setup Menu

].

3. Set [

Lead Set

] to [

12-Lead

], and set [

ECG Display

] to [

12-Lead

].

8-22

There are a total of 12 ECG waves and 1 rhythm wave displayed on the screen. The rhythm lead is ECG

Ⅰ before entering the 12-lead ECG screen.

The [

Filter

] mode is automatically switched to [

Diagnostic

] when the patient monitor accesses the 12-lead full-screen; the [

Filter

] mode resumes to the configuration before accessing the 12-lead full screen when the patient monitor exits the 12-lead full screen.

Passport 12m/Passport 17m Operator's Manual

8.10.1

Setting ECG Waveform Sequence

You can select the sequence of ECG waveforms on the 12-lead ECG screen and 12-lead ECG report.

To select the sequence of the ECG waveforms:

1. In the [

ECG Setup

] menu, select [

Others>>

] to enter the [

Other Setup Menu

].

2. Set [

Waveform Layout

] to [

Standard

] or [

Cabrera

].

[

Standard

]: the sequence is I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6.

[

Cabrera

]: the sequence is aVL, I, -aVR, II, aVF, III, V1, V2, V3, V4, V5, V6.

8.10.2

Extending the Rhythm Lead Waveform area

To extend the height of the rhythm lead waveform area:

1. In the [

ECG Setup

] menu, select [

Others>>

] to enter the [

Other Setup Menu

].

2. Set

[ECG Waveform Area

] to [

Extended

].

8.11

Resting 12-Lead ECG Analysis

The equipment incorporates the Glasgow algorithm, developed by the University of Glasgow, to provide an interpretation of the resting 12-lead ECG in all situations.

8.11.1

Accessing the 12-Lead Screen

To access the 12-lead screen:

1. In the [

ECG Setup

] menu, select [

Others>>

] to enter the [

Other Setup Menu

].

2. Set [

Lead Set

] to [

12-Lead

].

3. Set [

ECG Display

] to [

12-Lead

].

Passport 12m/Passport 17m Operator's Manual 8-23

The functions of the keys at the bottom of the 12-lead screen are as follows:

 [

Analyze

]: starts resting 12-lead analysis.

[

Patient demogr.

]: enters the patient information.

[

Setup

]: enters the 12-lead setup menu.

[

Report

]: prints the latest resting 12-lead report.

[

Exit

]: exits the 12-lead sceen.

8.11.2

Entering Patient Information

Some patient information may directly affect ECG analysis. Complete and correct patient information is helpful for accurate diagnosis and treatment of the patient. Enter patient information before taking an ECG measurement.

To enter the patient information, select [

Patient Demogr.

] from the 12-lead screen.

NOTE

Check that patient information is correct before resting 12-lead analysis.

We recommend using pediatric lead placement V4R, V1, V2, V4 - V6 if the patient is under 16 years of age.

Please record V4R using the V3 electrode. Also set [V3 Electrode Placement] to [V4R]. This is a normal practice for a patient of this age.

8-24 Passport 12m/Passport 17m Operator's Manual

8.11.3

12-Lead Setup

In the 12-lead screen, select [

Setup

] to enter the [

12-Lead Setup

] menu to change the settings related to 12-lead ECG analysis. In the [

12-Lead Setup

] menu, you can also select [

Report Setup>>

] to set the format and contents of the ECG reports.

12-lead Setup

Menu item Option Default Description

Filter Diagnostic, ST Diagnosti c

Baseline Drift

Removal

Tachy High

Brady Low

QTc Formula

On, Off

80 - 130

40 - 60

Hodges,

Bazett,

Fridericia,

Framingham

On

100

50

Hodges

Set filter mode.

Note:

The filter mode automatically switches to [

Diagnostic

] when the patient monitor accesses the 12-lead -screen and resumes to the original setting when the patient monitor exits the 12-lead screen.

Select whether the baseline drift removal (BDR) process or 0.05-Hz filter is used.

[

On

]: BDR is enabled. This process suppresses most baseline drift interference and also is able to preserve the fidelity of the ST-segment level.

[

Off

]: BDR is disabled and the 0.05-Hz filter is used.

NOTE:

BDR or 0.05-Hz selection applies to the displayed ECG, printed report, analyzed and stored data.

BDR introduces around 1-second delay. We recommend use of BDR except when the delay is unacceptable.

Both BDR and 0.05-Hz selections meet requirements of the 1990 American

Heart Association Recommendations for Standardization and Specifications in Automated Electrocardiography: Bandwidth and Signal Processing pertaining to lower-frequency response in electrocardiography.

Adjusts tachycardia threshold. Heart rates above the setting are labeled

Tachycardia.

Only applies to patients whose age exceeds 180 days.

Adjusts bradycardia threshold. Heart rates below the setting are labeled

Bradycardia.

Only applies to patients whose age exceeds 2191 days.

Selects QTc formula.

Hodges:

QTc

=

QT

+ 1 .

75 ×

(

HeartRate

− 60

)

Bazett:

QTc

=

QT

×

HeartRate

60

1

2

Waveform

Layout

Standard,

Cabrera

Fridericia:

QTc

=

QT

×

HeartRate

60

1

3

Framingham:

QTc

=

QT

+ 154 ×

60

HeartRate

Standard Select ECG lead sequence for displaying and printing.

[

Standard

]: the sequence is I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6;

[

Cabrera

]: the sequence is aVL, I, -aVR, II, aVF, III, V1, V2, V3, V4, V5, V6.

Passport 12m/Passport 17m Operator's Manual 8-25

12-lead Setup

Menu item Option

Report setup

Menu item Option

Report Format

12×1, 6×2,

3×4+1

Median

Complex

Measurements

Interpretation

Interpretation

Summary

On, Off

On, Off

On, Off

On, Off

Default Description

Default

3×4+1

Off

On

On

On

Description

Selects the format of the 12-lead ECG report.

[

12×1

]: ECG waveforms are displayed in 12 lines.

[

6×2

]: ECG waveforms are displayed in 6 lines and 2 columns.

[

3×4+1

]: ECG waveforms are displayed in 3 lines and 4 columns followed by the rhythm lead waveform.

Selects whether Median Complex is included on the 12-lead ECG report.

Median Complex displays a median complex waveform for each lead and a rhythm lead waveform of 10 seconds in 3x4+1 format. For each waveform, short vertical lines are used to mark the start of the P-wave and QRS complex, and the end of the P-wave, QRS complex, and T-wave.

Selects whether the measurement result is included on the 12-lead ECG report.

Measurement result includes Vent. Rate, PR Interval, QRS Duration, QT/QTc

Interval, and P/QRS/T Axes.

Selects whether diagnoses are included on the 12-lead ECG report.

Selects whether interpretation summary is included on the 12-lead ECG report.

Note: If the [

Interpretation

] option is not enabled, interpretation summary is not included on the report even if [

Interpretation Summary

] is enabled.

8

.11.4

Setting the 12-lead Order

To set whether to send the order of 12-lead interpretation report to the hospital information system while saving the report, follow this procedure:

1..

2..

elect [

Main Menu

]

[

User Maintenance

]

[

Others >>

]. e [

12-Lead Order

].

8.11.

5

Resting 12-lead ECG Analysis

Before 12-lead ECG interpretation, check that all electrodes are correctly connected to the lead wires and the ECG trunk cable is properly connected. Check that patient information is correct.

To start analyzing, select the [

Analyze

] key. The resting 12-lead analysis takes about 10 seconds. During this period, keep the patient still.

After analysis finishes, the following dialog-box pops up.

8-26 Passport 12m/Passport 17m Operator's Manual

Select [

Print Report

] to pint the resting 12-lead ECG report from the external printer.

You can also print the latest 12-lead ECG report by selecting [

Report

] from the 12-lead screen.

Refer to

12-Lead ECG Interpretive Program Physician’s Guide

(PN:

046-006360-00

) for details.

CAUTION

During the resting 12-lead ECG analysis, keep the patient still. Patient movement may cause misdiagnosis.

NOTE

Glasgow resting 12-lead ECG intepretation is applied to adult, pediatric and neonate.

During 12-lead ECG analaysis, 12-lead related settings are disabled.

Changing patient information, including the patient’s age, date of birth, gender, race, medication, class, or

V3 placement setting, may change diagnosis statement. You shall select the [Analyze] key to reanalyze the patient’s ECG before you print the latest 12-lead ECG report.

Passport 12m/Passport 17m Operator's Manual 8-27

8.11.

6

12-lead ECG Report

The format and contents of the 12-lead ECG report are configurable. Refer to

Report Setup

in

8.11.3 12-Lead Setup

for

details.

The following is a sample of the12-lead ECG report with default configuration.

1

2 3 4

8-28

5 6 7

1.

Patient information

2.

Time of resting 12-lead ECG analysis

3.

Measurements

4.

Diagnosis statement

5.

Waveform amplitude

6.

Paper speed

7.

Frequency range

8.

System software version/algorithm version

8

Passport 12m/Passport 17m Operator's Manual

8.12

Troubleshooting

This section lists the problems that might occur. If you encounter the problems when using the equipment or accessories, check the table below before requesting services. If the problem persists, contact your service personnel.

CAUTION

Never try to disassemble the equipment or supplied accessories. There are no internal user-serviceable parts.

Symptoms

Noisy ECG traces

Possible Cause

Loose or dry electrodes

Correction Action

Apply fresh and moist electrodes.

Defective electrode wires Replace wires if necessary.

Excessive Electro-surgical

Interference

Muscle Noise

Intermittent Signal

Excessive alarms: heart rate, lead fault

Low Amplitude ECG Signal

Patient cable or leads are routed too close to other electrical devices

Wrong ECG cable used

Inadequate skin preparation prior to application of electrode, tremors, tense subject, and/or poor electrode placement

Move the patient cable or leads away from the electrical device.

Use ESU-proof ECG cables. For details, refer to

section

35.1 ECG Accessories

.

Repeat skin preparation as described in section

8.3.1 Preparing the Patient and Placing the

Electrodes

and re-place the electrodes.

Apply fresh, moist electrodes.

Avoid muscular areas.

Check that the cables are properly connected. Connections not tight and/or properly secured

Electrodes dry or loose

Cable or lead wires damaged

Electrodes dry

Excessive patient movement or muscle tremor

Gain set too low

Electrodes dry / old

Skin improperly prepared

This could be the patient’s normal QRS complex

Electrode could be positioned over a bone or muscle mass

Repeat skin preparation as described in section

8.3.1 Preparing the Patient and Placing the

Electrodes

and apply fresh and moist electrodes.

Change cable and lead wires.

Repeat skin preparation as described in section

8.3.1 Preparing the Patient and Placing the

Electrodes

and apply fresh, moist electrodes.

Reposition the electrodes.

Replace fresh and moist electrodes if necessary.

Set the gain as required. For details, refer to section

8.5.3 Changing ECG Wave Settings

.

Apply fresh and moist electrodes.

Repeat skin preparation as described in section

8.3.1 Preparing the Patient and Placing the

Electrodes

.

Verify with another well-functioning monitor.

Move ECG patches away from the bone or muscle mass.

Passport 12m/Passport 17m Operator's Manual 8-29

Symptoms

No ECG Waveform

Base Line Wander

Possible Cause

Gain set too low

Lead wires and patient cable not fully or properly inserted

Cable or lead wires damaged

Patient moving excessively

Electrodes dry or loose

ECG Filter set to ST or Diagnostic mode

Correction Action

Set the gain as required. For details, refer to section

8.5.3 Changing ECG Wave Settings

.

Check that the leadwires and patient cables are properly connected.

Change cable and lead wires.

Secure leadwires and cable to patient.

Repeat skin preparation as described in section

8.3.1 Preparing the Patient and Placing the

Electrodes

and apply fresh and moist electrodes.

Set ECG Filter to “Monitor” mode.

8-30 Passport 12m/Passport 17m Operator's Manual

9

Monitoring Respiration (Resp)

9.1

Introduction

Impedance respiration is measured across the thorax. When the patient is breathing or ventilated, the volume of air changes in the lungs, resulting in impedance changes between the electrodes. Respiration rate (RR) is calculated from these impedance changes, and a respiration waveform appears on the monitor screen.

9.2

Safety Information

WARNING

When monitoring the patient’s respiration, do not use ESU-proof ECG cables.

If you do not set the detection level for the respiration correctly in manual detection mode, it may not be possible for the monitor to detect apnea. If you set the detection level too low, the monitor is more likely to detect cardiac activity, and to falsely interpret cardiac activity as respiratory activity in the case of apnea.

The respiration measurement does not recognize the cause of apneas. It only indicates an alarm if no breath is detected when a pre-configured time has elapsed since the last detected breath. Therefore, it cannot be used for diagnostic purpose.

If operating under conditions according to the EMC Standard EN 60601-1-2 (Radiated Immunity 3V/m), field strengths above 1V/m may cause erroneous measurements at various frequencies. Therefore it is recommended to avoid the use of electrically radiating equipment in close proximity to the respiration measurement unit.

9.3

Understanding the Resp Display

2 1

1. Gain

2. Resp lead label

3

4

3. Respiration rate

4. RR source

By selecting the waveform area or parameter area, you can enter the [

Resp Setup

] menu.

NOTE

Respiration monitoring is not for use on the patients who are very active, as this will cause false alarms.

Passport 12m/Passport 17m Operator's Manual 9-1

9.4

Placing Resp Electrodes

As the skin is a poor conductor of electricity, preparing the skin is necessary for a good respiration signal. You can refer to

the ECG section for how to prepare the skin. For details, refer to section

8.3.1 Preparing the Patient and Placing the

Electrodes

.

As the respiration measurement adopts the standard ECG electrode placement, you can use different ECG cables (3-lead,

5-lead or 12-lead). Since the respiration signal is measured between two ECG electrodes, if a standard ECG electrode placement is applied, the two electrodes should be RA and LA of ECG Lead I, or RA and LL of ECG Lead II.

NOTE

To optimize the respiration waveform, place the RA and LA electrodes horizontally when monitoring respiration with ECG Lead I; place the RA and LL electrodes diagonally when monitoring respiration with

ECG Lead II.

9-2

Lead I Lead II

Passport 12m/Passport 17m Operator's Manual

9.4.1

Optimizing Lead Placement for Resp

If you want to measure Resp and you are already measuring ECG, you may need to optimize the placement of the two electrodes between which Resp will be measured. Repositioning ECG electrodes from standard positions results in changes in the ECG waveform and may influence ST and arrhythmia interpretation.

9.4.2

Cardiac Overlay

Cardiac activity that affects the Resp waveform is called cardiac overlay. It happens when the Resp electrodes pick up impedance changes caused by the rhythmic blood flow. Correct electrodes placement can help to reduce cardiac overlay: avoid the liver area and the ventricles of the heart in the line between the respiratory electrodes. This is particularly important for neonates.

9.4.3

Abdominal Breathing

Some patients with restricted movement breathe mainly abdominally. In these cases, you may need to place the left leg electrode on the left abdomen at the point of maximum abdominal expansion to optimise the respiratory wave.

9.4.4

Lateral Chest Expansion

In clinical applications, some patients (especially neonates) expand their chests laterally, causing a negative intrathoracic pressure. In these cases, it is better to place the two respiration electrodes in the right midaxillary and the left lateral chest areas at the patient’s maximum point of the breathing movement to optimise the respiratory waveform.

9.5

Choosing the Respiration Lead

In the [

Resp Setup

] menu, set [

Resp Lead

] to [

I

], [

II

] or [

Auto

].

9.6

Changing the Apnea Alarm Delay

The apnea alarm is a high-level alarm used to detect apneas. You can set the apnea alarm delay time after which the monitor alarms if the patient stops breathing.

To change the apnea alarm delay:

1. In the [

Resp Setup

] menu, select [

Apnea Delay

].

2. Select the appropriate setting.

The [

Apnea Delay

] of Resp, CO

2

, AG, and RM module keeps consistent with each other.

Passport 12m/Passport 17m Operator's Manual 9-3

9.7

Changing Resp Detection Mode

In the [

Resp Setup

] menu, select [

Detection Mode

] and then select [

Auto

] or [

Manual

].

 In auto detection mode, the monitor adjusts the detection level automatically, depending on the wave height and the presence of cardiac artifact. Note that in auto detection mode, the detection level (a dotted line) is not displayed on the waveform.

Use auto detection mode for situations where:

The respiration rate is not close to the heart rate.

Breathing is spontaneous, with or without continuous positive airway pressure (CPAP).

Patients are ventilated, except patients with intermittent mandatory ventilation (IMV).

In manual detection mode, you adjust the dotted detection level line to the desired level by selecting [

Upper Line

] or [

Lower Line

] and then selecting or beside them. Once set, the detection level will not adapt automatically to different respiration depths. It is important to remember that if the depth of breathing changes, you may need to change the detection level.

Use manual detection mode for situations where:

The respiration rate and the heart rate are close.

Patients have intermittent mandatory ventilation.

Respiration is weak. Try repositioning the electrodes to improve the signal.

In Auto Detection Mode, if you are monitoring Resp and ECG is switched off, the monitor cannot compare the ECG and

Resp rates to detect cardiac overlay. The respiration detection level is automatically set higher to prevent the detection of cardiac overlay as respiration.

In Manual Detection Mode, cardiac overlay can in certain situations trigger the respiration counter. This may lead to a false indication of a high respiration or an undetected apnea condition. If you suspect that cardiac overlay is being registered as breathing activity, raise the detection level above the zone of cardiac overlay. If the Resp wave is so small that raising the detection level is not possible, you may need to optimize the electrode placement as described in

section

9.4.4 Lateral Chest Expansion.

9.8

Changing Resp Wave Settings

WARNING

When monitoring in manual detection mode, make sure to check the respiration detection level after you have increased or decreased the size of the respiration wave.

In the [

Resp Setup

] menu, you can:

Select [

Gain

] and then select an appropriate setting. The bigger the gain is, the larger the wave amplitude is.

Select [

Sweep

] and then select an appropriate setting. The faster the wave sweeps, the wider the wave is.

9-4 Passport 12m/Passport 17m Operator's Manual

9.9

Setting Respiration Rate (RR) Source

To set RR source:

1. Enter the [

Resp Setup

] menu.

2. Select [

RR Source

] and then select a source or [

Auto

] from the dropdown list.

The dropdown list displays the currently available RR source. When you select [

Auto

], the system will automatically select the RR source according to the priority. When the current RR source does not have valid measurement, the system will automatically switch the [

RR Source

] to [

Auto

]. RR source switches back to impedance respiration if you press the

[

Alarm Reset

] hardkey on the monitor’s front panel during an apnea alarm.

The priority of RR source is (from high to low): CO

2

measurement, RM measurement and impedance respiration measurement.

The [

RR Source

] settings of Resp, CO

2

, AG and RM module are linked.

The RR source options and description are shown in the table below.

Option

Auto

CO

2

RM

ECG

Description

RR source is automatically selected according to the priority.

RR source is from CO

2

measurement.

RR source is from RM measurement.

RR source is from impedance respiration measurement.

9.10

Setting Alarm Properties

Select [

Alarm Setup >>

] from the [

Resp Setup

] menu. In the pop-up menu, you can set alarm properties for this parameter.

Passport 12m/Passport 17m Operator's Manual 9-5

FOR YOUR NOTES

9-6 Passport 12m/Passport 17m Operator's Manual

10

Monitoring PR

10.1

Introduction

The pulse numeric counts the arterial pulsations that result from the mechanical activity of the heart. You can display a pulse from any measured SpO

2 or any arterial pressure (see

chapter

14 Monitoring IBP

). The displayed pulse numeric is

color-coded to match its source.

1. PR: detected beats per minute.

2. PR Source

1

2

NOTE

A functional tester or SpO

2

simulator can be used to determine pulse rate accuracy.

10.2

Setting the PR Source

The current pulse source is displayed in the PR parameter area. The pulse rate chosen as pulse source:

 is monitored as system pulse and generates alarms when you select PR as the active alarm source;

 is stored in the monitor’s database and reviewed in the graphic/tabular trends; in trend graphs, as the PR curve is in the same color with the PR source , it is unlikely to distinguish the PR source ;

 is sent via the network to the CMS, if available.

To set which pulse rate as PR source:

1. Enter the [

SpO2 Setup]

menu.

2. Select [

PR Source

] and then select a label or [

Auto

] from the pop-up menu.

The pop-up menu displays the currently available PR sources from top to bottom by priority. When you select [

Auto

], the system will automatically select the first option as the PR source from the pop-up menu. When the current PR source is unavailable, the system will automatically switch [

PR Source

] to [

Auto

]. When you select [

IBP

], the system will automatically select the first pressure label as the PR source from the pop-up menu.

Passport 12m/Passport 17m Operator's Manual 10-1

10.3

Selecting the Active Alarm Source

In most cases the HR and pulse numerics are identical. In order to avoid simultaneous alarms on HR and PR, the monitor uses either HR or PR as its active alarm source. To change the alarm source, select [

Alm Source

] in the [

ECG Setup

] or

[

SpO2 Setup

] menu and then select either:

 [

HR

]: The monitor will use the HR as the alarm source for HR/pulse.

[

PR

]: The monitor will use the PR as the alarm source for HR/pulse.

[

Auto

]: If the [

Alm Source

] is set to [

Auto

], the monitor will use the heart rate from the ECG measurement as the alarm source whenever the ECG measurement is switched on and a valid heart rate is available. If the heart rate becomes unavailable, for example if leads becomes disconnected, and a pulse source is switch on and available, the monitor will automatically switch to Pulse as the alarm source. When the Leads Off condition is corrected, the monitor will automatically switch back to the heart rate as the alarm source.

10.4

QRS Tone

When PR is used as the alarm source, the PR source will be used as a source for the QRS tone. You can change the QRS volume by adjusting [

Beat Vol

] in the [

SpO2 Setup

] menu. When a valid SpO

2

value exists, the system will adjust the pitch tone of QRS volume according to the SpO

2

value.

10-2 Passport 12m/Passport 17m Operator's Manual

11

Monitoring SpO

2

11.1

Introduction

SpO

2

monitoring is a non-invasive technique used to measure the amount of oxygenated haemoglobin and pulse rate by measuring the absorption of selected wavelengths of light. The light generated in the probe passes through the tissue and is converted into electrical signals by the photodetector in the probe. The SpO

2

module processes the electrical signal and displays a waveform and digital values for SpO

2

and pulse rate.

This device is calibrated to display functional oxygen saturation. It provides the following measurements:

1 2

3

4

6

1

2

3

4 5

1.

Pleth waveform (Pleth/Plethb): visual indication of patient’s pulse. The waveform is not normalized.

2.

Oxygen saturation of arterial blood (SpO

2

/SpO

2 b): percentage of oxygenated hemoglobin in relation to the sum of oxyhemoglobin and deoxyhemoglobin. SpO

2

measurement is obtained through the MPM module, and SpO

2 b measurement is obtained through the SpO

2

module.

3.

Perfusion index (PI): gives the numerical value for the pulsatile portion of the measured signal caused by arterial pulsation. PI is an indicator of the pulsatile strength. You can also use it to assess the quality of SpO

2

measurement.

Above 1 is optimal.

Between 0.3 and 1 is acceptable.

Below 0.3 indicates low perfusion. When PI is below 0.3, a question mark (?) is displayed to the right of the

SpO

2

value, indicating that the SpO

2

value may be inaccurate. Reposition the SpO

2

sensor or find a better site. If low perfusion persists, choose another method to measure oxygen saturation if possible.

PI is available for Masimo SpO

2

module. For Masimo SpO

2

module, PI value can be displayed under the PR value in larger characters if [

PI Zoom

] is enabled.

4.

Perfusion indicator: the pulsatile portion of the measured signal caused by arterial pulsation.

5.

SpO

2

difference (ΔSpO

2

):

Δ

SpO

2

=

SpO

2

SpO

2 b

.

6.

Pulse rate (derived from the pleth wave): detected pulsations per minute.

If you need to measure SpO

2

and SpO

2 b, select the same type of modules. Otherwise, the SpO

2

module for SpO

2 b is closed automatically. For example, if MPM module configuring Nellcor SpO

2

and Masimo SpO

2 module are applied simultaneously, Masimo SpO

2

module is closed automatically.

Passport 12m/Passport 17m Operator's Manual 11-1

NOTE

A functional tester or SpO

2

simulator cannot be used to assess the accuracy of a SpO

2 module or a SpO

2 sensor.

11.2

Safety

WARNING

Use only SpO

2

sensors specified in this manual. Follow the SpO

2

sensor’s instructions for use and adhere to all warnings and cautions.

When a trend toward patient deoxygenation is indicated, blood samples should be analyzed by a laboratory co-oximeter to completely understand the patient’s condition.

The equipment is not intended for use in an MRI environment.

Prolonged continuous monitoring may increase the risk of undesirable changes in skin characteristics, such as irritation, reddening, blistering or burns. Inspect the sensor site every two hours and move the sensor if the skin quality changes. Change the application site every four hours. For neonates, or patients with poor peripheral blood circulation or sensitive skin, inspect the sensor site more frequently.

11.3

Identifying SpO

2

Modules

To identify which SpO

2

module is incorporated into your MPM, T1, N1, or SpO

2

module, see the color of the connector and the company logo located at the right upper corner. The color of the cable connector matches the company as shown below:

Masimo SpO

2

module: a purple connector with a logo of Masimo SET.

Nellcor SpO

2

module: a grey connector with a logo of Nellcor.

The connectors for these SpO

2

sensors are mutually exclusive.

11.4

Applying the Sensor

To apply the sensor:

1. Select an appropriate sensor according to the module type, patient category and weight.

2. Remove colored nail polish from the application site.

3. Apply the sensor to the patient.

4. Select an appropriate adapter cable according to the connector type and plug this cable into the SpO

2

connector.

5. Connect the sensor cable to the adapter cable.

11-2 Passport 12m/Passport 17m Operator's Manual

WARNING

If the sensor is too tight because the application site is too large or becomes too large due to edema, excessive pressure for prolonged periods may result in venous congestion distal from the application site, leading to interstitial edema and tissue ischemia.

11.5

Changing SpO

2

Settings

11.5.1

Accessing SpO2 Menus

By selecting the SpO

2

parameter window or waveform area, you can access the [

SpO2 Setup

] or [

SpO2b Setup

] menu.

11.5.2

Adjusting the Desat Alarm

The desat alarm is a high level alarm notifying you of potentially life threatening drops in oxygen saturation. Select

[

Alarm Setup >>

] from the [

SpO2 Setup

] or [

SpO2b Setup

] menu. From the pop-up menu, you can set low alarm limit, alarm switch, and alarm recording for [

Desat

] or [

Desatb

]. When the SpO

2

value is below the desat alarm limit and desat alarm switch is set on, the message [

SpO2 Desat] or

[

SpO2b Desat]

is displayed.

11.5.3

Setting SpO2 Sensitivity

For Masimo SpO

2

module, you can set [

Sensitivity

] to [

Normal

] or [

Maximum

] in the [

SpO2 Setup

] or [

SpO2b Setup

] menu. When the [

Sensitivity

] is set to [

Maximum

], the monitor is more sensitive to minor signals. When monitoring critically ill patients whose pulsations are very weak, it is strongly recommended that the sensitivity is set to [

Maximum

].

When monitoring neonatal or non-critically ill patients who tend to move a lot, noise or invalid signals may be caused. In this case, it is recommended that the sensitivity is set to [

Normal

] so that the interference caused by motion can be filtered and therefore the measurement stability can be ensured. The settings of sensitivity in the [

SpO2 Setup

] and

[

SpO2b Setup

] menus are linked.

11.5.4

Changing Averaging Time

The SpO

2

value displayed on the monitor screen is the average of data collected within a specific time. The shorter the averaging time is, the quicker the monitor responds to changes in the patient’s oxygen saturation level. Contrarily, the longer the averaging time is, the slower the monitor responds to changes in the patient’s oxygen saturation level, but the measurement accuracy will be improved. For critically ill patients, selecting shorter averaging time will help understanding the patient’s state.

To set the averaging time for Masimo SpO

2

module:

1. Select [

Averaging

] in the [

SpO2 Setup

] or [

SpO2b Setup

] menu.

2. Then select [

2-4 s

], [

4-6 s

], [

8 s

], [

10 s

], [

12 s

], [

14 s

] or [

16 s

].

Passport 12m/Passport 17m Operator's Manual 11-3

11.5.5

Monitoring SpO2 and NIBP Simultaneously

When monitoring SpO

2

and NIBP on the same limb simultaneously, you can switch [

NIBP Simul

] on in the [

SpO2 Setup

] or [

SpO2b Setup

] menu to lock the SpO

2

alarm status until the NIBP measurement ends. If you switch [

NIBP Simul

] off, low perfusion caused by NIBP measurement may lead to inaccurate SpO

2

readings and therefore cause false physiological alarms.

11.5.6

Sat-Seconds Alarm Management

With traditional alarm management, high and low alarm limits are set for monitoring oxygen saturation. During monitoring, as soon as an alarm limit is violated, an audible alarm immediately sounds. When the patient % SpO

2 fluctuates near an alarm limit, the alarm sounds each time the limit is violated. Such frequent alarms can be distracting.

Nellcor’s Sat-Seconds alarm management technique is used to reduce these nuisance alarms.

The Sat-Seconds feature is available with the Nellcor SpO

2

module to decrease the likelihood of false alarms caused by motion artifacts. To set the Sat-Seconds limit, select [

Sat-Seconds

] in the [

SpO2 Setup

] menu and then select the appropriate setting.

With Sat-Seconds alarm management, high and low alarm limits are set in the same way as traditional alarm management. A Sat-Seconds limit is also set. The Sat-Seconds limit controls the amount of time that SpO

2

saturation may be outside the set limits before an alarm sounds. The method of calculation is as follows: the number of percentage points that the SpO

2

saturation falls outside the alarm limit is multiplied by the number of seconds that it remains outside the limit. This can be stated as the equation:

Sat-Seconds= Points × Seconds

Only when the Sat-Seconds limit is reached, the monitor gives a Sat-Seconds alarm. For example, the figure below demonstrates the alarm response time with a Sat-Seconds limit set at 50 and a low SpO

2

limit set at 90%. In this example, the patient % SpO

2 drops to 88% (2 points) and remains there for 2 seconds. Then it drops to 86% (4 points) for 3 seconds, and then to 84% (6 points) for 6 seconds. The resulting Sat-Seconds are:

% SpO

2

Total Sat-Seconds=

Seconds

2=

3=

6=

Sat-Seconds

4

12

36

52

After approximately 10.9 seconds, a Sat-Second alarm would sound, because the limit of 50 Sat-Seconds would have been exceeded.

11-4 Passport 12m/Passport 17m Operator's Manual

% SpO

2

Seconds

Saturation levels may fluctuate rather than remaining steady for a period of several seconds. Often, the patient % SpO

2 may fluctuate above and below an alarm limit, re-entering the non-alarm range several times. During such fluctuation, the monitor integrates the number of %SpO

2

points, both positive and negative, until either the Sat-Seconds limit is reached, or the patient%SpO

2

re-enters the non-alarm range and remains there.

NOTE

The “SpO

2

Too Low” or “SpO

2

Too High” alarm is presented in the case that SpO

2 value violates the alarm limits for 3 times within one minute even if the setting of Sat-Seconds is not reached.

11.5.7

Changing the Speed of the Pleth/Plethb Wave

In the [

SpO2 Setup

] or [

SpO2b Setup

] menu, select [

Sweep

] and then select the appropriate setting. The faster the waveform sweeps, the wider the waveform is.

11.5.8

Zooming PI Value

For Masimo SpO

2

module, you can display PI value in larger characters for better view. To zoom in the display of PI value, set [

PI Zoom

] to [

Yes

] from the [

SpO2 Setup

] menu.

11.5.9

Setting the Alarm Level for SpO2 Sensor Off Alarm

Select [

Alarm Setup >>

] from the [

User Maintenance

] menu. You can set the [

SpO2 SensorOff Lev.

] in the pop-up menu.

Passport 12m/Passport 17m Operator's Manual 11-5

11.5.10

Setting the SpO2 Tone Mode

Select [

Others >>

] from the [

User Maintenance

] menu. In the pop-up menu, you can set [

SpO2 Tone

] as [

Mode 1

] or

[

Mode 2

].

WARNING

The same SpO

2

tone mode shall be used for the same monitors in a single area.

11.6

Measurement Limitations

If you doubt the SpO

2 measurement, check the patient’s vital signs first. Then check the patient monitor and SpO

2

sensor.

The following factors may influence the accuracy of measurement:

Ambient light

Physical movement (patient and imposed motion)

Diagnostic testing

Low perfusion

Electromagnetic interference, such as MRI environment

Electrosurgical units

Dysfunctional haemoglobin, such as carboxyhemoglobin (COHb)and methemoglobin (MetHb)

Presence of certain dyes, such as methylene and indigo carmine

Inappropriate positioning of the SpO

2

sensor, or use of incorrect SpO

2 sensor

 Drop of arterial blood flow to immeaurable level caused by shock, anemia, low temperature or vasoconstrictor.

11-6 Passport 12m/Passport 17m Operator's Manual

11.7

Masimo Information

Masimo Patents

This posting serves as notice under 35 U.S.C.

§

287(a) for Masimo patents: http://www.masimo.com/patents.htm.

 No Implied License

Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized sensors or cables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.

11.8

Nellcor Information

 Nellcor Patents

This posting serves as notice under 35 U.S.C.

§

287(a) for Covidien patents: http://www.covidien.com/patents.

No Implied License

Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized replacement parts which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.

Passport 12m/Passport 17m Operator's Manual 11-7

11.9

Troubleshooting

This section lists the problems that might occur. If you encounter the problems when using the equipment or accessories, check the table below before requesting for services. If the problem persists, contact your service personnel.

CAUTION

Never try to disassemble the equipment or supplied accessories. There are no internal user-serviceable parts.

Symptoms

Dashes “- -” display in place of numerics.

Do not see SpO

2

parameter tiles in display.

Possible Cause

Measurement is invalid. the application site if necessary.

Parameter not configured to display. Switch the SpO

2 monitoring function on as

described in section

3.13.1

Switching the

Parameters On/Off.

Unable to obtain SpO

2

reading Patient has poor perfusion

Sensor not on patient

Cables loose/not connected

Ambient light

Correction Action

Check that the sensor is properly applied. Change

Switch limbs/notify physician

Check if the [

SPO2 Sensor Off

] alarm is reported.

If so, reapply the sensor.

If not, contact the service personnel.

Check connections, switch cable

Check if the [

SpO2 Too Much Light

] alarm is reported. If so, move the sensor to a place with lower level of ambient light or cover the sensor to minimize the ambient light.

No SpO

2

waveform

Low amplitude SpO

2 signal

Waveform not selected to display

Cable or sensor not plugged in

SpO

2

sensor on same limb as cuff

Patient has poor perfusion

Switch the SpO

2

monitoring function on as

described in section

3.13.1 Switching the

Parameters On/Off

Check that the cable is properly connected and sensor securely applied.

Check that the sensor is properly applied. Change the application site if necessary.

Change the application site.

11-8 Passport 12m/Passport 17m Operator's Manual

12

Monitoring NIBP

12.1

Introduction

The MPM, T1, and N1 use the oscillometric method for measuring the non-invasive blood pressure (NIBP). This measurement can be used for adults, pediatrics and neonates.

Automatic non-invasive blood pressure monitoring uses the oscillometric method of measurement. To understand how this method works, we’ll compare it to the auscultative method. With auscultation, the clinician listens to the blood pressure and determines the systolic and diastolic pressures. The mean pressure can then be calculated with reference to these pressures as long as the arterial pressure curve is normal.

Since the monitor cannot hear the blood pressure, it measures cuff pressure oscillation amplitudes. Oscillations are caused by blood pressure pulses against the cuff. The oscillation with the greatest amplitude is the mean pressure. This is the most accurate parameter measured by the oscillometric method. Once the mean pressure is determined, the systolic and diastolic pressures are calculated with reference to the mean.

Simply stated, auscultation measures systolic and diastolic pressures and the mean pressure is calculated. The oscillometric method measures the mean pressure and determines the systolic and diastolic pressures.

As specified by IEC 80601-2-30, NIBP measurement can be performed during electro-surgery and discharge of defibrillator.

NIBP diagnostic significance must be decided by the clinician who performs the measurement.

NOTE

Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultatory method or an intra-arterial blood pressure measurement device, within the limits prescribed by the American National Standard, Manual, electronic, or automated sphygmomanometers.

Passport 12m/Passport 17m Operator's Manual 12-1

12.2

Safety

WARNING

Be sure to select the correct patient category setting for your patient before measurement. Do not apply the higher adult settings for pediatric or neonatal patients. Otherwise it may present a safety hazard.

Do not measure NIBP on patients with sickle-cell disease or on the limb where skin damage has occurred or is expected.

Use clinical judgment to determine whether to perform frequent unattended blood pressure measurements on patients with severe blood clotting disorders because of the risk of hematoma in the limb fitted with the cuff.

Do not use the NIBP cuff on a limb with an intravenous infusion or arterial catheter in place. This could cause tissue damage around the catheter when the infusion is slowed or blocked during cuff inflation.

NIBP reading can be affected by the measurement site, the position of the PATIENT, exercise, or the patient's physiologic condition. If you doubt the NIBP measurements, determine the patient’s vital signs by alternative means and then verify that the monitor is working correctly.

Do not apply cuff on the arm on the side of a mastectomy.

Do not modify or replace connectors of the NIBP air hose except with mindray-approved connectors. Use neonatal and infant cuffs with CM1901 hoses only. Use pediatric/adult cuffs with CM1903 hoses only.

Never connect intra-arterial or intra-venous lines, or any other incompatible connectors to the NIBP hose.

This can cause serious injury or death.

Avoid placing cuff on patient in a manner that can lead to a hose becoming kinked (hose kinking may cause inaccurate readings).

12.3

Measurement Limitations

Measurements are impossible with heart rate extremes of less than 40bpm or greater than 240bpm, or if the patient is on a heart-lung machine.

The measurement may be inaccurate or impossible:

If a regular arterial pressure pulse is hard to detect

 With excessive and continuous patient movement such as shivering or convulsions

 With cardiac arrhythmias

 Rapid blood pressure changes

 Severe shock or hypothermia that reduces blood flow to the peripheries

 Obesity, where a thick layer of fat surrounding a limb dampens the oscillations coming from the artery

12-2 Passport 12m/Passport 17m Operator's Manual

12.4

Measurement Methods

There are three methods of measuring NIBP:

 Manual: measurement on demand.

Auto: continually repeated measurements at set intervals.

STAT: continual rapid series of measurements over a five minute period, then return to the previous mode.

12.5

Setting Up the NIBP Measurement

12.5.1

Preparing the Patient

In normal use, perform NIBP measurement on a patient who is in the following position:

 Comfortably seated

 Legs uncrossed

 Feet flat on the floor

 Back, arm and feet supported

Middle of the cuff at the level of the right atrium of the heart

NOTE

It is recommended that the patient relaxes as much as possible before performing measurement and that the patient does not talk during measurement.

To avoid unnecessary motion artifact, it is recommended to have the patient sit quietly for several minutes before taking the measurement.

The operator shall not touch the cuff or tubing during NIBP measurement.

Other factors that have been shown to result in an overestimation of blood pressure are labored breathing, full bladder, pain etc.

12.5.2

Preparing to Measure NIBP

To prepare for the NIBP measurement:

1. Power on the monitor.

2. Verify that the patient category is correct. If not, select the [

Patient Setup

] QuickKey

[

Patient Demographics

]

[

Patient Cat.

] and set the patient category to [

Adu

], [

Ped

] or [

Neo

]

3. Plug the air tubing into the NIBP connector on the MPM module or T1, or N1.

4. Select a correct sized cuff and then wrap it directly over the patient’s skin as follows:

Determine the patient’s limb circumference.

Select an appropriate cuff by referring to the limb circumference marked on the cuff. The width of the cuff should be 40% (50% for neonates) of the limb circumference, or 2/3 of the upper arm’s length. The inflatable part of the cuff should be long enough to encircle at least 50% to 80% of the limb.

Apply the cuff to the patient’s upper arm or leg and make sure the Φ marking on the cuff matches the artery location. The cuff should fit snugly, but with enough room for two fingers to be placed between the cuff and

Passport 12m/Passport 17m Operator's Manual 12-3

the patient’s arm (on adults), and loosely on neonates with little or no air present within the cuff. Otherwise, it may cause discoloration, and ischemia of the extremities. Make sure that the cuff edge falls within the marked range. If it does not, use a larger or smaller cuff that will fit better.

5. Connect the cuff to the air tubing. Avoid compression or restriction of pressure tubes. Air must pass unrestricted through the tubing.

WARNING

Continuous cuff pressure due to connection tubing kinking may cause blood flow interference and resulting harmful injury to the patient.

NOTE

Equipment use is restricted to one patient at a time.

12.5.3

Starting and Stopping Measurements

Select the [

NIBP Measure

] QuickKey and you can start the desired measurement from the pop-up menu. You can select

[

Stop All

] QuickKey to stop all NIBP measurements. You can also start and stop measurements by using the [

NIBP

] hardkey on either the monitor’s front panel or the MPM module.

12.5.4

Correcting the Measurement if Limb is not at Heart Level

The cuff should be applied to a limb at the same level as the patient’s heart. If the limb is not at the heart level, to the displayed value:

Add 0.75 mmHg (0.10 kPa) for each centimetre higher, or

Deduct 0.75 mmHg (0.10 kPa) for each centimeter lower.

12.5.5

Enabling NIBP Auto Cycling and Setting the Interval

To enable the NIBP auto cycling and set the interval:

1. Select the NIBP parameter window to enter the [

NIBP Setup

] menu.

2. Select [

Interval

] and then select a desired time interval. Selecting [

Manual

] switches to manual mode.

3. Start a measurement manually. The monitor will then automatically repeat NIBP measurements at the set time interval.

12.5.6

Enabling Measurement on Clock

In auto measurement mode, if the clock is enabled, the monitor automatically synchronizes NIBP automatic measurements with the real time clock. For example, if [

Interval

] is set to [

20 min]

, and you start NIBP auto measurement at 14: 03, the next measurement will be taken at 14: 20, and then at 14:40, 15:00, and so on.

12-4 Passport 12m/Passport 17m Operator's Manual

To enable measurement on clock:

1.

Select the NIBP parameter window to enter the [

NIBP Setup

] menu.

2.

Set [

Clock

] to [

On

].

NOTE

The measurement on clock feature is only available for patient monitors supporting Advanced ECG algorithm.

Measurement on clock is effective only when NIBP Interval is set to [5min] or an option greater than 5 min.

12.5.7

Starting a STAT Measurement

To start a STAT measurement:

1. Select the NIBP parameter window to enter the [

NIBP Setup

] menu.

2. Select [

NIBP STAT

].

The STAT mode initiates 5 minutes of continuous, sequential, automatic NIBP measurements.

WARNING

Continuous non-invasive blood pressure measurements may cause purpura, ischemia and neuropathy in the limb with the cuff. Inspect the application site regularly to ensure skin quality and inspect the cuffed limb for normal color, warmth and sensitivity. If any abnormity occurs, move the cuff to another site or stop the blood pressure measurements immediately.

12.6

Understanding the NIBP Numerics

The NIBP display shows numerics only as below. Your display may be configured to look slightly different.

1

2

3

4

5

1. Time of last measurement

2. Time remaining to next measurement

3. Measurement mode

4. Unit of pressure: mmHg or kPa

5. Systolic pressure

6. Diastolic pressure

6 7

7. Mean pressure obtained after the measurement and cuff pressure obtained during the measurement

Passport 12m/Passport 17m Operator's Manual 12-5

If the NIBP measurement exceeds the measurement range, “---” will be displayed.

12.7

Changing NIBP Settings

By selecting the NIBP parameter window, you can enter the [

NIBP Setup

] menu.

12.7.1

Setting the Initial Cuff Inflation Pressure

You can set the initial cuff inflation pressure manually. In the [

NIBP Setup

] menu, select [

Initial Pressure

] and then select the appropriate setting.

NOTE

For known hypertensive patients, you need to set initial cuff pressure to a higher value to reduce the measurement time.

12.7.2

Setting NIBP Alarm Properties

Select [

Alarm Setup >>

] from the [

NIBP Setup

] menu. You can set the alarm properties for this parameter in the pop-up menu.

12.7.3

Displaying NIBP List

Select [

Screens

] QuickKey

[

Screen Setup

]. You can set [

NIBP List

] to be displayed at the bottom area of the screen, where the most recent NIBP measurements display. Then, multiple sets of most recent NIBP measurements will be displayed. And the displayed PR is derived from NIBP.

You can not display NIBP list in some screens such as the big numerics screen.

12.7.4

Setting the Pressure Unit

Select [

Unit Setup >>

] from the [

User Maintenance

] menu. In the pop-up menu, select [

Press.

Unit

] and then select

[

mmHg

] or [

kPa

].

12.7.5

Switching On NIBP End Tone

The monitor can issue a reminder tone at the completion of NIBP measurement. The NIBP end tone is off by default. You can switch it on by accessing the [

NIBP Setup

] menu.

12-6 Passport 12m/Passport 17m Operator's Manual

12.8

Assisting Venous Puncture

You can use the NIBP cuff to cause sub-diastolic pressure to block the venous blood vessel and therefore help venous puncture.

1. Select [

VeniPuncture >>

] from the [

NIBP Setup

] menu. In the pop-up menu, verify that the [

Cuff Press.

] value is appropriate. Change it if necessary.

2. Select [

VeniPuncture

].

3. Puncture vein and draw blood sample.

4. Select the [

NIBP

] hardkey on the monitor’s front panel, or the [

Stop All

] QuickKey to deflate the cuff. The cuff deflates automatically after a set time if you do not deflate it.

During measurement, the NIBP display shows the inflation pressure of the cuff and the remaining time in venous puncture mode.

Passport 12m/Passport 17m Operator's Manual 12-7

FOR YOUR NOTES

12-8 Passport 12m/Passport 17m Operator's Manual

13

Monitoring Temp

13.1

Introduction

The equipment is used to monitor skin temperature and core temperature. You can simultaneously monitor two temperature sites using the MPM, the T1, or the N1.

13.2

Safety

WARNING

Verify that the probe detection program works correctly before monitoring. Remove the temperature probe cable from the T1 or T2 connector, and the monitor can display the message [T1 Sensor Off] or [T2 Sensor

Off] and give alarm tones correctly.

13.3

Taking a Temp Measurement

To take a Temp measurement:

1. Select an appropriate probe for your patient according to patient type and measuring site.

2. If you are using a disposable probe, connect the probe to the temperature cable.

3 Plug the probe or temperature cable to the temperature connector.

4. Attach the probe to the patient correctly.

5. Check that the alarm settings are appropriate for this patient.

13.4

Understanding the Temp Display

The temperature monitoring is displayed on the monitor as three numerics: T1, T2 and TD. By selecting this area, you can enter the [

Alarm Setup

] menu.

Passport 12m/Passport 17m Operator's Manual 13-1

13.5

Setting the Temperature Unit

Select [

Unit Setup >>

] from the [

User Maintenance

] menu. In the pop-up menu, select [

Temp Unit]

and then select [

ºC

] or [

ºF

].

13-2 Passport 12m/Passport 17m Operator's Manual

14

Monitoring IBP

14.1

Introduction

You can measure invasive blood pressure (IBP) using the MPM, T1, N1, or the pressure plug-in module. The Passport 12m and Passport 17m patient monitor can monitor up to 8 invasive blood pressures and displays the systolic, diastolic and mean pressures and a waveform for each pressure.

1 2

3

1. Setup key

2. Zero key

3. IBP cable connector

14.2

Safety

WARNING

Use only pressure transducers specified in this manual. Never reuse disposable pressure transducers.

Make sure that the applied parts never contact other conductive parts.

To reduce the hazard of burns during high-frequency surgical procedure, ensure that the monitor’s cables and transducers never come into contact with the high-frequency surgical units.

When using accessories, their operating temperature should be taken into consideration. For details, refer to instructions for use of accessories.

The neutral electrode of the electro-surgery unit (ESU) shall properly contact the patient. Otherwise, burns may result.

Passport 12m/Passport 17m Operator's Manual 14-1

14.3

Zeroing the Transducer

To avoid inaccurate pressure readings, the monitor requires a valid zero. Zero the transducer in accordance with your hospital policy (at least once per day).

Zero whenever:

A new transducer or adapter cable is used.

You reconnect the transducer cable to the monitor.

The monitor restarts.

You doubt the readings.

To zero the transducer:

1. Turn off the stopcock to the patient.

Pressure transducer

3-way stopcock

Pressure adapter cable

Monitor

2. Vent the transducer to the atmospheric pressure by turning on the stopcock to the air.

3. Press the hardkey on the module, or, in the setup menu for the pressure (e.g. Art), select [

Art Zero >>

]

[

Zero

]. During zero calibration, the [

Zero

] button appears dimmed. It recovers after the zero calibration is completed. To zero all IBP channels, select [

Zero IBP

] hotkey, and then select [

Zero All Channels

] in the pop-up menu.

4. After the zero calibration is completed, close the stopcock to the air and open the stopcock to the patient.

NOTE

Your hospital policy may recommend that the ICP transducer is zeroed less frequently than other transducers.

14-2 Passport 12m/Passport 17m Operator's Manual

14.4

Setting Up the Pressure Measurement

To set up the pressure measurement:

1. Plug the pressure cable into the IBP connector.

2. Prepare the flush solution.

3. Flush the system to exhaust all air from the tubing. Ensure that the transducer and stopcocks are free of air bubbles.

WARNING

If air bubbles appear in the tubing system, flush the system with the infusion solution again. Air bubbles may lead to incorrect pressure readings.

4. Connect the pressure line to the patient catheter.

5. Position the transducer so that it is level with the heart, approximately at the level of the midaxillary line.

6. Select the appropriate label.

7. Zero the transducer. After a successful zeroing, turn off the stopcock to the air and turn on the stopcock to the patient.

Heparinized fluid bag

Pressure transduce

Valve

Flushing set

Pressure line

Stopcock

Adapter cable for using disposable transducers

Monitor

WARNING

If measuring intracranial pressure (ICP) with a sitting patient, level the transducer with the top of the patient’s ear. Incorrect leveling may give incorrect values.

Passport 12m/Passport 17m Operator's Manual 14-3

14.5

Understanding the IBP Display

The IBP measurement is displayed on the monitor as a waveform and numeric pressures. The figure below shows the waveform and numerics for the Art pressure. For different pressures, this display may be slightly different.

6

1

1. Pressure label

2. Waveform

3. Systolic pressure

2

3 4 5

4. Diastolic pressure

5. Mean pressure

6. PPV measurement

For some pressures, the parameter window may show the mean pressure only. For different pressures, their defaults unit may be different. If the Art and ICP pressures are measured simultaneously, the ICP parameter area will display numeric

CPP, which is obtained by subtracting ICP from the Art mean.

14.6

Changing IBP Settings

14.6.1

Changing a Pressure for Monitoring

To change a pressure for monitoring:

1.

Select the pressure you want to change to enter its setup menu.

In the menu, there is a figure showing the current IBP measurement connector.

14-4 Passport 12m/Passport 17m Operator's Manual

Label

PA

Ao

UAP

BAP

FAP

Art

P1 to P4

NOTE

2. Select [

Label

] and then select your desired label from the list. Labels already used cannot be selected.

Description

Pulmonary artery pressure

Aortic pressure

Umbilical arterial pressure

Brachial arterial pressure

Femoral arterial pressure

Arterial blood pressure

Non-specific pressure label

Label

CVP

LAP

RAP

ICP

UVP

LV

Description

Central venous pressure

Left atrial pressure

Right atrial pressure

Intracranial pressure

Umbilical venous pressure

Left ventricular pressure

When two pressures are detected having the same label, the monitor automatically changes one pressure label to a currently unused one.

14.6.2

Setting Alarm Properties

Select [

Alarm Setup >>

] from the parameter setup menu. You can set alarm properties for this parameter in the pop-up menu.

14.6.3

Changing Averaging Time

The IBP value displayed on the monitor screen is the average of data collected within a specific time. The shorter the averaging time is, the quicker the monitor responds to changes in the patient’s blood pressure. Contrarily, the longer the averaging time is, the slower the monitor responds to changes in the patient’s blood pressure, but the measurement accuracy will be improved. For critically ill patients, selecting shorter averaging time will help understanding the patient’s state.

To set the averaging time, in the parameter setup menu, select [

Sensitivity

] and then select [

High

], [

Med

] or [

Low

], the corresponding averaging time is about 1 s, 8 s and 12 s respectively.

14.6.4

Setting Up the IBP Wave

In the setup menu for the pressure, you can:

Select [

Sweep

] and then select the appropriate setting. The faster the wave sweeps, the wider the wave is.

Select [

Scale

] and then select the appropriate setting. If [

Auto

] is selected, the size of the pressure’s waveform will be adjusted automatically.

Select [

Filter

] and then select the desired option.

14.6.5

Setting the Pressure Unit

Select [

Unit Setup >>

] from the [

User Maintenance

] menu. In the pop-up menu, select [

Press. Unit

] and then select

[

mmHg

] or [

kPa

]. Select [

CVP Unit

] and then select [

mmHg

], [

cmH

2

O

] or [

kPa

].

Passport 12m/Passport 17m Operator's Manual 14-5

14.6.6

Enabling PPV Measurement and Setting PPV Source

PPV indicates pulse pressure variation. To enable PPV measurement, set [

PPV Measurement

] to [

On

].

You can select PPV source by enabling the PPV measurement.

WARNING

This monitor can calculate PPV from beat-to-beat values of any arterial pulsatile pressure. The circumstances under which the calculation of a PPV value is clinically meaningful, appropriate and reliable must be determined by a physician.

The clinical value of the derived PPV information must be determined by a physician. According to recent scientific literature, the clinical relevance of PPV information is restricted to sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia.

PPV calculation may lead to inaccurate values in the following situations:

at respiration rates below 8 rpm

during ventilation with tidal volumes lower than 8 ml/kg

for patients with acute right ventricular dysfunction (“cor pulmonale”).

The PPV measurement has been validated only for adult patients.

14.7

Overlapping IBP Waveforms

The IBP waveforms can be displayed together. To combine IBP waveforms:

1. Select [

Main Menu

]

[

Screen Setup>>

]

[

Screen Layout>>

] to access the [

Screens

] window.

2. Select the [

Screen Setup

] tab.

3. In Area A, select [

IBP Overlap

] from the drop-down list, and then select the IBP waves to be overlapped on the left

side of the same line. Refer to section

3.10 Setting the Screen

for Area A.

4. Select to save the setting and exit the window. The main screen will display the overlapped IBP waves.

14-6

Selecting the overlapped IBP waveforms on the main screen opens the [

Overlapping Waveform Setup

] menu, where you can:

Set [

Left Scale

] and

[Right Scale

] and then set the scales for the overlapped waveforms. The left scale is for Art, LV,

Ao, FAP, BAP, UAP, and the arterial waveforms of P1~P4; the right scale is for CVP, ICP, LAP, RAP, UVP, and the venous waveforms of P1~P4.

Set [

CVP Scale

] individually if the CVP waveform is combined and CVP unit is different from IBP unit.

Passport 12m/Passport 17m Operator's Manual

 Set [

PA Scale

] individually if PA waveform is combined.

Set [

Gridlines

] to [

On

] or [

Off

] to show gridlines or not in the overlapped waveform area.

Select [

Sweep

] and then set the sweep speed for the overlapped waveforms.

Select [

Filter

] and then set the filter for the overlapped waveforms.

You can also change above settings from corresponding IBP setup menu.

Note

The CVP scale changes simultaneously with the right scale. The unit of CVP scale is consistent with CVP parameter unit.

14.8

Measuring PAWP

Pulmonary Artery Wedge Pressure (PAWP) values, used to assess cardiac function, are affected by fluid status, myocardial contractility, and valve and pulmonary circulation integrity.

Obtain the measurement by introducing a balloon-tipped pulmonary artery flotation catheter into the pulmonary artery.

When the catheter is in one of the smaller pulmonary arteries, the inflated balloon occludes the artery allowing the monitor to record changes in the intrathoracic pressures that occur throughout the respiration cycle.

The pulmonary wedge pressure is the left ventricular end diastolic pressure when the airway pressure and valve function are normal. The most accurate PAWP values are obtained at the end of the respiration cycle when the intrathoracic pressure is fairly constant and the artifact caused by respiration is minimal.

WARNING

PAWP monitoring is not intended for neonatal patients.

14.8.1

Preparing to Measure PAWP

To prepare for a PAWP measurement:

1. Prepare the same accessories as in the C.O. measurement. Connect the parts such as catheter, syringe, etc.

following the C.O. measurement steps and use the balloon inflation port. Refer to section

15.4 Setting Up the C.O.

Measurement

for more information.

2. Connect the PAWP cable into the IBP connector. Since PAWP is measured on PA, selecting [

PA

] as the IBP label is recommended.

3. Select the PA parameter window or waveform area to enter its setup menu. Then, select [

PAWP

] to enter the PAWP measurement window. You can also enter the PAWP measurement window from the P1-P4 parameter window.

Passport 12m/Passport 17m Operator's Manual 14-7

NOTE

After entering the PAWP measurement window, the monitor will turn off the PA alarm automatically.

14.8.2

Setting Up the PAWP Measurement

To set up the PAWP measurement:

1. Wedge the flotation catheter into the pulmonary artery. Then inflate the balloon and pay attention to the PA waveform changes on the screen.

14-8

2. After obtaining a stable PAWP waveform, press the [

Freeze

] key to freeze the waveform and deflate the balloon.

3. You can adjust the PAWP scale to an appropriate position by adjusting or beside the [

Adjust

] button.

Press the [

Confirm

] key to save one PAWP measurement.

4. If you need to start a new measurement, select [

Next Measure

].

Passport 12m/Passport 17m Operator's Manual

WARNING

Prolonged inflation can cause pulmonary hemorrhage, infarction or both. Inflate the balloon for the minimum time necessary to get an accurate measurement.

If the PAWP is greater than the PA (systolic), deflate the balloon and report the incident in accordance with hospital policy. Because the pulmonary artery could be accidentally ruptured, and the PAWP value derived will not reflect the patient’s hemodynamic state, but will merely reflect the pressure in the catheter or balloon.

14.8.3

Understanding the PAWP Setup Menu

Select [

Setup

] to enter the [

PAWP Setup

] menu. In this menu, you can:

 Select an ECG lead wave as the first reference wave.

Select a respiration wave as the second reference wave.

Select a sweep speed for the displayed waveform.

Change the size of the PA waveform by adjusting the scale height.

Passport 12m/Passport 17m Operator's Manual 14-9

14.9

Troubleshooting

This section lists the problems that might occur. If you encounter the problems when using the equipment or accessories, check the table below before requesting for services. If the problem persists, contact your service personnel.

CAUTION

Never try to disassemble the equipment or supplied accessories. There are no internal user-serviceable parts.

Symptoms

Damped invasive waveform

Possible Cause

Air bubbles in tubing

IBP not displayed/no IBP waveform

Dashes “- -” display in place of numerics.

Abnormally high or low readings

Kinked catheter

Blood in tubing

Improper setup

Cable not plugged in

Transducer not connected.

Stopcock turned improperly.

Transducer not zeroed

The measured result is invalid or out of range.

IBP might be set to non-pulsatile labels like

CVP, LA, RA, and ICP.

Transducer too High or too Low.

Correction Action

Eliminate air from tubing as described in

section

14.4 Setting Up the Pressure

Measurement

.

Change the position of catheter.

Pressurize the solution bag to 300 mmHg. For details, refer to the instructions for use of the solution bag.

Check display setup in monitor setup.

Check that the cables are properly connected.

Check that the transducer is properly connected.

Check that the stopcock is turned to the correct position.

Check and zero the transducer as described in

section

14.3 Zeroing the Transducer

Change to a pulsatile label.

Unable to Zero

PAWP button disabled

Stopcock not open to atmosphere.

One IBP channel must be labeled PA

Adjust the position of the transducer and make sure that it is level with the heart, approximately at the level of the midaxillary line.

Zero the transducer as described in section

14.3 Zeroing the Transducer

.

Check the transducer and make sure the stopcock is turned to the air.

Label an IBP channel as PA. (Also Label an IBP channel as P1/P2/P3/P4, it will automatically change to PA)

14-10 Passport 12m/Passport 17m Operator's Manual

15

Monitoring Cardiac Output

15.1

Introduction

The cardiac output (C.O.) measurement invasively measures cardiac output and other hemodynamic parameters using the right heart (atria) thermodilution method. A cold solution of known volume and temperature is injected into the right atrium through the proximal port of a pulmonary artery (PA) catheter. The cold solution mixes with the blood in the right ventricle and the change in blood temperature is measured with a thermistor at the distal end of the catheter in the pulmonary artery. The temperature change is displayed as a curve in the C.O. split screen, and the monitor calculates the C.O. value from this curve. The C.O. value is inversely proportional to the area under the curve. As cardiac output varies continuously, a series of measurements must be carried out to achieve a reliable C.O. average value. Always use the average of multiple thermodilution measurements for therapy decisions. The monitor is capable of storing 6 measurements.

Setup key

C.O. start key

C.O. cable connector

WARNING

C.O. monitoring is restricted to adult patients only.

Passport 12m/Passport 17m Operator's Manual 15-1

15.2

Understanding the C.O. Display

The C.O. measurement is displayed on the monitor as numeric C.O., C.I. and TB in the C.O. parameter window as shown below. To enter the [

C.O. Setup

] menu, select the C.O. parameter window.

2

1

1. Cardiac output

2. Time at which the C.O. average is calculated

3. Cardiac index

4. Blood temperature

3

4

15.3

Influencing Factors

The factors that affect cardiac output are:

 temperature of injectate solution,

 volume of injectate solution,

 patient’s baseline blood temperature,

 patient’s inspiratory/expiratory cycle,

 placement of catheter with relation to proximity of lung field,

 the catheter itself,

 the patient rhythm and hemodynamic status, and

 any other rapid IV solutions which are infused while the C.O. measurement is being performed.

Following are some technical suggestions to obtain accurate C.O.:

 Injectate solution must be cooler than the patient’s blood.

Inject solution rapidly and smoothly.

Inject at end of expiration.

15-2 Passport 12m/Passport 17m Operator's Manual

15.4

Setting Up the C.O. Measurement

WARNING

Use only accessories specified in this manual. Make sure that the accessories never come into contact with conductive parts.

To set up the C.O. measurement:

1. Connect the C.O. cable to the C.O. connector.

2. Interconnect the C.O. module, catheter and syringe as shown below. Make sure that:

The module is securely inserted.

The PA catheter is in place in the patient.

The C.O. cable is properly connected to the module.

C.O. module

C.O. cable

Injectate

PA catheter

Syringe

Stopcock

Hole

In-line probe

Balloon inflation port

Ice water

NOTE

The above picture illustrates when using a TI sensor (PN 6000-10-02079). The connection may be different if other TI sensors are used.

3. Select the C.O. parameter window to enter the [

C.O. Setup

] menu. Check if the height and weight are appropriate for your patient. Change if necessary.

Passport 12m/Passport 17m Operator's Manual 15-3

4. In the [

C.O. Setup

] menu:

Check that the correct computation constant is entered. Refer to the Instruction for Use of pulmonary artery catheter to determine the [

Comp. Const

] according to the entered injectate volume and temperature. To change the computation constant, select [

Comp. Const

] and then enter the correct value. When a new catheter is used, the computation constant should be adjusted in accordance with the manufacturer’s instructions for use.

Set the [

Auto TI

] to [

Manual

] or [

Auto

]. If you select [

Auto

], the system automatically detects the injectate temperature, and the [

Manual TI (°C)

] is disabled. If you select [

Manual

], you need to enter the injectate temperature at [

Manual TI (°C)

] manually.

Set the [

Measuring Mode

] to [

Auto

] or [

Manual

]. In [

Auto

] mode, the monitor automatically takes the C.O. measurement after establishing a baseline blood temperature. In [

Manual

] mode, you need to click the [

Start

] button in the C.O. measurements window when the monitor is ready for new C.O. measurement.

5. Select [

Enter C.O. Screen

] to enter the C.O. measurements window.

15-4

A

B

C

A. Currently measured numeric

B. Currently measured C.O. curve

C. Prompt message area

D. Buttons

E. Averaged values

F. Measurement windows

D

6. Proceed as follows.

In [

Manual

] measure mode, select the [

Start

] button and then inject the solution quickly when you see the message [

Ready for new set of measurement

],. As shown in the figure above, during the measurement, the currently measured thermodilution curve is displayed. At the end of the measurement, the thermodilution curve is transferred to one of the 6 measurement windows and the monitor prompts you to wait for a certain period of time before starting a new measurement.

F

E

Passport 12m/Passport 17m Operator's Manual

In [

Auto

] measure mode, the monitor consecutively takes C.O. measurements automatically without the need for pressing the [

Start

] button between two measurements. A new thermodilution measurement is possible as soon as the message [

Inject now!

] is displayed on the screen. The monitor automatically detects further thermodilution measurements.

7. Consecutively take 3 to 5 single measurements as instructed by Step 6.

A maximum of 6 measurements can be stored. If you perform more than six measurements without rejecting any, the oldest will automatically be deleted when a seventh curve is stored. Select from the 6 measurement curves and the system will automatically calculate and display the averaged C.O. and C.I. values. Then select the [

Accept Average

] button to accept and store the averaged values.

When injecting, the stopcock to the PA catheter is open and the stopcock to the injectate solution is closed. After the measurement is completed, turn off the stopcock to the PA catheter and turn on the stopcock to the injectate solution, and then draw the injectate solution into the injectate syringe.

In the buttons area, you can:

Select [

Start

] to start a C.O. measurement.

 Select [

Stop

] to stop the current measurement.

 Select [

Cancel

] during a measurement to cancel the measurement.

 Select [

Record

] to print out the curves selected for average calculation, numerics and averaged values by the recorder.

Select [

Setup >>

] to access the [

C.O. Setup

] menu.

Select [

Calc >>

]

[

Hemodynamic >>

] to access the [

Hemodynamic Calculation

] menu.

The system can automatically adjust the X-axis scale range to 30 s or 60 s and Y-axis scale range to 0.5°C, 1.0°C, or 2.0°C.

CAUTION

Starting a measurement without blood temperature being stable may cause measurement failure.

NOTE

During a cardiac output measurement, blood temperature alarms are inactive.

Please refer to the Instructions for Use of pulmonary artery catheter of the monitor to determine the [Comp.

Const] and the volume of injectate.

Passport 12m/Passport 17m Operator's Manual 15-5

15.5

Measuring the Blood Temperature

As shown below, the blood temperature is measured with a temperature sensor at the distal end of the catheter in the pulmonary artery. During C.O. measurements, blood temperature alarms are suppressed to avoid false alarms. They will automatically recover as soon as the C.O. measurements are completed.

Pulmonary artery

PA catheter

Balloon

Right atrium

Sensor

Right ventricle

15.6

Changing C.O. Settings

15.6.1

Setting the Temperature Unit

Select [

Unit Setup >>

] from the [

User Maintenance

] menu. In the pop-up menu, select [

Temp Unit

] to toggle between

[

°C

] and [

°F

].

15.6.2

Setting Alarm Properties

Select [

Alarm Setup >>

] from the [

C.O. Setup

] menu. You can set alarm properties for this parameter in the pop-up menu.

15-6 Passport 12m/Passport 17m Operator's Manual

16

Monitoring CCO/SvO

2

16.1

Introduction

The Edwards Vigilance II® monitors, Vigileo TM monitors, and EV1000 monitors measure continuous cardiac output (CCO), mixed venous oxygen saturation (SvO

2

), central venous oxygen saturation (ScvO

2

) etc. They also calculate hemodynamic and oxygenation parameters. The Passport 12m and Passport 17m monitors can be connected to the Vigilance II® monitor/Vigileo TM monitor / EV1000 monitor and can display, store, and review the measured and calculated parameter values from these monitors. The Passport 12m and Passport 17m monitors can also give alarms of these measured parameters. You must set alarm on/off, alarm limits, alarm level, and alarm record separately on this monitor. The alarm is

[

Off

] by default.

CCO setup menu Signal output setup menu

Indicator

CCO/SvO

2

cable connector

16.2

Safety

WARNING

The Vigilance II® monitor, Vigileo TM monitor, and EV1000 monitor are manufactured by Edwards

Lifesciences. This company provides the technology of measuring and calculating the relevant parameters.

We only provide the connection between this monitor and Vigilance II® monitor/Vigileo TM monitor/ EV1000 monitor.

If you have any doubts about the operation and maintenance of the Vigilance II® monitor/Vigileo TM monitor/

EV1000 monitor, please read the Operator’s Manuals for these monitors or contact Edwards Lifesciences

(www.edwards.com) directly.

Fully observe the Vigilance II® monitor/Vigileo TM monitor/ EV1000 monitor Operator’s Manuals to configure settings and to connect the monitor to the patient.

Passport 12m/Passport 17m Operator's Manual 16-1

NOTE

When the Vigilance II® monitor, the Vigileo

TM

monitor, or the EV1000 monitor is disconnected from the

Passport 12m or the Passport 17m monitor, a disconnection alarm is announced. In addition, the CCO numeric area and the SvO

2

/ScvO

2

numeric area disappear from the monitor’s screen.To clear the disconnection alarm, please see the ScvO

2

section of

D.2 Technical Alarm Messages

.

16.3

Automatic Communication Detection

The relevant parameter window is not displayed on the screen if this monitor detects communication failure between the CCO/SvO

2

module and Vigilance II® monitor/Vigileo

TM

monitor/ EV1000 monitor.

16.4

Connecting the Device

16.4.1

Connecting the Vigilance II® Monitor

The following figure shows how to connect this monitor to the Vigilance II® monitor through cables.

The following figure shows the rear housing of the Vigilance II® monitor.

16-2

Serial ports ECG signal input port

Analog signal input ports

Passport 12m/Passport 17m Operator's Manual

To connect the Vigilance II® monitor:

1. Connect CN1 with the CCO/SvO

2

connector on the patient monitor.

2. Insert the ECG signal end into the ECG signal input port marked monitor.

on the rear housing of the Vigilance II®

3. Insert the MAP signal end into the analog signal input port 1 marked , the CVP signal end into port 2 marked

, and SPO

2

signal end into port 3 marked respectively on the rear housing of the Vigilance II® monitor.

4. Insert UART into either of the serial ports (marked

5. Set the Vigilance II® monitor as follows:

) on the rear housing of the Vigilance II® monitor

 Access the [

Serial Port Setup

] menu.

Set [

Device

] to [

IFMou

t], [

Baud Rate

] to [

19200

], [

Parity

] to [

None

], [

Stop Bits

] to [

1

], [

Data Bits

] to [

8

], and

[

Flow Control

] to [

2 s

].

Access the [

Analog Input Setup

] menu.

For port 1, set [

Parameter

] to [

MAP

], [

Voltage Range

] to [

0-5 v

], [

Full Scale Range

] to

500 mmHg

(66.7 kPa),

[

Simulated High Value

] to

500 mmHg

(66.7 kPa), and [

Simulated Low Value

] to

0 mmHg

(0.0 kPa).

For port 2, set [

Parameter

] to [

CVP

], [

Voltage Range

] to [

0-5 v

], [

Full Scale Range

] to

100 mmHg

(13.3 kPa),

[

Simulated High Value

] to

100 mmHg

(13.3 kPa), and [

Simulated Low Value

] to

0 mmHg

(0.0 kPa).

For port 3, set [

Parameter

] to [

SaO

2

], [

Voltage Range

] to [

0-10 v

], [

Full Scale Range

] to [

100%

], [

Simulated

High Value

] to [

100%

], and [

Simulated Low Value

] to [

0%

].

Refer to the Vigilance II® Operator’s Manual for the operation of the monitor.

WARNING

Calibrate the Vigilance II® monitor before monitoring. Refer to the Vigilance II® Operator’s Manual for the calibration instructions.

NOTE

For the Vigilance II® monitor, [Flow Control] must be set to 2 seconds.

Passport 12m/Passport 17m Operator's Manual 16-3

16.4.2

Connecting the Vigileo

TM

Monitor

The following figure shows how to connect this monitor to the Vigileo

TM

monitor through cables.

The following figure shows the rear housing of the Vigileo

TM

monitor.

Patient Optical Module(OM)

Analog input

Analog output

Serial port

16-4

To connect the Vigileo

TM

monitor:

1. Connect CN1 with the CCO/SvO

2

connector on the monitor.

2. Insert the CVP signal end into the analog signal input port on the rear housing of the Vigileo TM monitor.

3. Insert UART into the serial port on the rear housing of the Vigileo TM monitor.

4. Set the Vigileo TM monitor as follows:

Access the [

Serial Port Setup

] menu.

Set [

Device

] to [

IFMou

t], [

Baud Rate

] to [

19200

], [

Parity

] to [

None

], [

Stop Bits

] to [

1

], [

Data Bits

] to [

8

], and

[

Flow Control

] to [

2 seconds

].

Access the [

Analog Input Port Setup

] menu.

Set [

Parameter

] to [

CVP

], [

Voltage Range

] to [

0-5 v

], [

Full Scale Range

] to

100 mmHg

(13.3 kPa), [

High Value

] to

100 mmHg

(13.3 kPa), and [

Simulated Low Value

] to

0 mmHg

(0.0 kPa).

Simulated

Refer to the Vigileo

TM

Operator’s Manual for the operation of the monitor.

Passport 12m/Passport 17m Operator's Manual

WARNING

Calibrate the Vigileo TM monitor before monitoring. Refer to the Vigileo TM Operator’s Manual for the calibration instructions.

NOTE

For the Vigileo TM monitor, [Flow Control] must be set to 2 seconds.

16.4.3

Connecting the EV1000 Monitor

The following figure shows how to connect this monitor to the EV1000 monitor through cables.

The following figure shows the rear housing of the EV1000 monitor.

Passport 12m/Passport 17m Operator's Manual

Serial port

16-5

To connect the EV1000 monitor:

1. Connect CN1 with the CCO/SvO

2

connector on the patient monitor.

2. Insert UART into the serial port on the rear housing of the EV1000 monitor.

3. On the main screen of the EV1000 monitor, select the button to access the [

Settings

] menu.

4. Select [

Monitor Settings

] and then select [

Serial Port Setup

] to access the [

Serial Port Setup

] menu.

5. Set [

Device

] to [

IFMou

t], [

Baud Rate

] to [

19200

], [

Parity

] to [

None

], [

Stop Bits

] to [

1

], [

Data Bits

] to [

8

], and [

Flow

Control

] to [

2 s

].

Refer to the EV1000 Operator’s Manual for the operation of the monitor.

WARNING

Calibrate the EV1000 monitor before monitoring. Refer to the EV1000 Operator’s Manual for the calibration instructions.

16.5

Understanding CCO Parameters

CCI

C.O.

C.I.

EDV

EDVI

SV

SVI

When the monitor is connected to the Vigilance II® monitor/Vigileo

TM

monitor/EV1000 monitor, select the CCO parameter window

[

Hemodynamic Parameters >>

] to view the hemodynamic parameters for evaluation of the patient’s hemodynamic status.

16.5.1

Hemodynamic Parameters for Vigilance II® Monitor

Abbreviation Unit Full Spelling Measurement Range

CCO L/min continuous cardiac output 1 to 20 L/min

Resolution

0.1

SVR

SVRI

RVEF

TB

ESV

ESVI

L/min/m 2

L/min

L/min/m 2 ml ml/m

2 ml ml/m

2

DS/cm 5 kPa-s/l

DS·m 2 /cm 5 kPa-s-m

2

/l

%

°C

°F ml ml/m

2 continuous cardiac index cardiac output cardiac index end diastolic volume end diastolic volume index stroke volume stroke volume index systemic vascular resistance systemic vascular resistance index right ventricular ejection fraction blood temperature end systolic volume end systolic volume index

0 to 20 L/min/m 2

1 to 20 L/min

0 to 20 L/min/m 2

40 to 800 ml

20 to 400 ml/m

2

0 to 300 ml

0 to 200 ml/m

2

0 to 3000 DS/cm 5

0 to 300 kPa-s/l

0 to 6000 DS·m 2 /cm 5

0 to 600 kPa-s-m

2

/l

10 to 60%

25 to 45

77 to 113

10 to 700 ml

5 to 400 ml/m

2

1

1

1

0.1

0.1

0.1

1

1

0.1

1

0.1

1

1

1

0.1

0.1

16-6 Passport 12m/Passport 17m Operator's Manual

Abbreviation Unit

CVP

MAP

HR cmH

2

O, kPa or mmHg central venous pressure mmHg or kPa mean arterial pressure rpm heart rate

16.5.2

Hemodynamic Parameters for Vigileo

TM

Monitor

Abbreviation Unit Full Spelling

0 to 100 mmHg

0 to 500 mmHg

30 to 250 bpm

CCO

CCI

SV

SVI

L/min

L/min/m

2 ml ml/m

2

Full Spelling

continuous cardiac output continuous cardiac index stroke volume stroke volume index

Measurement Range

Measurement Range

1 to 20 L/min

0 to 20 L/min/m

2

0 to 300 ml

0 to 200 ml/m

2

Resolution

1

1

1

Resolution

0.1

0.1

1

1

SVV

SVR

SVRI

CVP

%

DS/cm 5 kPa-s/l

DS·m

2

/cm

5 kPa-s-m2/l cmH

2

O, kPa or mmHg stroke volume variation systemic vascular resistance systemic vascular resistance index central venous pressure

0 to 99%

0 to 3000 DS/cm 5

0 to 300 kPa-s/l

0 to 6000 DS·m

2

/cm

5

0 to 600 kPa-s-m

2

/l

0 to 100 mmHg

0.1

1

0.1

1

0.1

1

16.5.3

Hemodynamic Parameters for EV1000 Monitor

When the EV1000 monitor is in the FloTrac or ClearSight mode, you can view parameter values for CCO, CCI, SV, SVI, SVV,

SVR, SVRI, CVP, and MAP on the Passport 12m and Passport 17m monitors’ screen.

When the EV1000 monitor is in the VolumeView mode, you can view parameter values for CCO, CCI, C.O, C.I., SV , SVI, GEF,

CFI, GEDV, GEDI, ITBV , ITBI, SVV, SVR , SVRI, EVLW , ELWI, PVPI, TB, CVP, and MAP on the Passport 12m and Passport 17m monitors’ screen.

Abbreviation Unit Full Spelling Measurement Range Resolution

Output

Contractility

C.O.

C.I.

SV

SVI

GEF

CFI

Preload

Volume

CCO

CCI

GEDV

GEDI

L/min continuous output cardiac

L/min/m

L/min

2 continuous index cardiac output

L/min/m

2 cardiac index cardiac

1 to 20 L/min

0 to 20 L/min/m 2

1 to 20 L/min

0 to 20 L/min/m

2 ml ml/m 2

%

1/min ml ml/m

2 stroke volume stroke volume index

0 to 300 ml

0 to 200 ml/m 2 global ejection fraction 1 to 99% cardiac function index 1 to 15 L/min global end diastolic volume global end diastolic volume index

40 to 4800 ml

80 to 2400 ml/m

2

0.1

0.1

1

1

0.1

0.1

1

0.1

1

1

Passport 12m/Passport 17m Operator's Manual 16-7

Afterload

Organ

Function

Abbreviation Unit

ITBV

ITBI

SVV

CVP

SVR

SVRI

MAP

EVLW

ELWI

PVPI

TB

Full Spelling Measurement Range

ml ml/m 2 intra-thoracic blood volume

50 to 6000 ml intra-thoracic blood volume index

100 to 3000 ml/ m stroke volume variation 0 to 99%

2

% cmH

2

O, kPa, or mmHg none

°C

°F central venous pressure 0 to 300 mmHg

DS/cm kPa-s/l

DS·m

2

5

/cm

5

0 to 3000 DS/cm 5 resistance

0 to 300 kPa-s/l systemic vascular

2

/cm

5 resistance index

0 to 600 kPa-s-m

2

/l kPa-s-m

2

/l mmHg or kPa mean arterial pressure 0 to 300 mmHg ml ml/kg extravascular water extravascular water index lung

10 to 5000 ml lung

0 to 50 ml/kg pulmonary vascular permeability iIndex

0.1 to 9.9 blood temperature

15 to 45°C

59 to 113°F

Resolution

1

1

0.1

1

1

0.1

1

0.1

1

1

0.1

0.1

0.1

0.1

16.6

Understanding the CCO Display

The parameter area displays the parameter measurements from the Vigilance II® monitor/Vigileo

TM

monitor/EV1000 monitor. You can select the desired parameters to be displayed.

The parameters display differently in the continuous measurement mode and intermittent measurement mode (for the

Vigilance II® monitor and Vigileo TM monitor only). You can select the desired parameters to be displayed in both modes.

For the configuration of the parameters to be displayed, see section

16.7.2 Selecting the Displayed Parameters

.

16-8 Passport 12m/Passport 17m Operator's Manual

16.7

Changing CCO Settings

16.7.1

Selecting Vascular Resistance Unit

To select vascular resistance unit:

1. Access the [

CCO Setup

] menu.

2. Select [

SVR Unit

] and toggle between [

DS/cm5

] and [

kPa-s/l

].

16.7.2

Selecting the Displayed Parameters

16.7.2.1

Selecting the Displayed Parameters for Vigilance II Monitor and Vigileo

TM

Monitor

To select the parameter to be displayed:

1. Access the [

CCO Setup

] menu.

2. Select [

Select Parameter >>

].

3. Select the parameters to be displayed from the pop-up menu.

16.7.2.2

Selecting the Displayed Parameters for EV1000 Monitor

To select the parameter to be displayed:

1.

Access the [

CCO Setup

] menu.

2.

Select [

Select Parameter >>

].

3.

Set[

Parameter Display

] to [

Absolute

] or [

Indexed

].

4.

Select the desired secondary parameters. You can select up to three secondary parameters.

16.7.3

Checking the C.O. Measurements

When the monitor is connected to the Vigilance II® monitor, you can check the C.O. measurements in the intermittent measurement mode.

To check the C.O. measurements:

1. Access the [

CCO Setup

] menu.

2. Select [

C.O. Measurements >>

].

Passport 12m/Passport 17m Operator's Manual 16-9

16.7.4

Setting Signal Output

When the monitor is connected to the Vigilance II® monitor:

This monitor outputs analog signals for the Vigilance II® monitor. You can select [

Signal Output Setup >>

] from the

[

CCO Setup

] menu to set the source of MAP signals. You can also select [

Simulated High Value

] or [

Simulated

Low Value

] to provide simulated high value or low value signals for calibrating the Vigilance II® monitor. Refer to the Vigilance II® Operator’s Manual for the calibration of the monitor.

When the monitor is connected to the Vigileo

TM

monitor:

Select [

Signal Output Setup >>

] from the [

CCO Setup

] menu. In the pop-up menu, you can select [

Simulated

High Value

] or [

Simulated Low Value

] to provide simulated high value or low value signals for calibrating the

Vigileo

TM

monitor. Refer to the Vigileo

TM

Operator’s Manual for the calibration of the monitor.

16.7.5

Selecting Alarm Properties

You can select [

Alarm Setup >>

] from the [

CCO Setup

] menu to set the parameter alarm properties.

WARNING

Because the alarm limits of the relevant measured parameters can be set on this monitor, the alarms of these parameters may be different from those on the Vigilance II® monitor/Vigileo TM monitor/EV1000 monitor. Please pay special attention to the alarms on the Vigilance II® monitor /Vigileo TM monitor/EV1000 monitor.

The alarm of the relevant measured parameters on this monitor is Off by default. Please pay special attention to the alarms on the Vigilance II® monitor/Vigileo TM monitor/EV1000 monitor.

The following table lists the alarm limits for hemodynamic parameters that come from Vigilance II

®

Monitor, Vigileo TM

Monitor, and EV1000 Monitor.

Alarm Limit Range Step

CCO High

CCO Low

(Low limit+0.1) to 25 L/min

0.3 to(high limit-0.1)L/min

0.1 L/min

Applicable For

Vigilance II

®

Monitor ,

Vigileo TM Monitor, and

EV1000 Monitor

CCI High (Low limit+0.1) to 15 L/min/m 2

0.1 L/min/m

2

CCI Low 0.1 to(high limit-0.1)L/min/m

2

EDV High (Low limit+10)to 800 ml

Vigilance II

®

Monitor

10 ml

EDV Low 0 to (high limit-10)ml

EDVI High (Low limit+10) to 400 ml/m

2

10 ml/m

2

EDVI Low

SVR High

SVR Low

SVRI High

0 to (high limit-10)ml/m 2

(Low limit+20) to 5000 DS/cm

5 or (low limit+2) to 500 kPa-s/l

0 to (high limit-20)DS/cm 5 or 0 to (high limit-2)kPa-s/l

(Low limit+50) to 9950 DS·m

2

/cm

5 or (low limit+5) to 995 kPa-s-m 2 /l

20 DS/cm

5 or2 kPa-s/l

50 DS·m

2

/cm or 5 kPa-s-m

5

2 /l

Vigilance II

®

Monitor ,

Vigileo TM Monitor, and

EV1000 Monitor

16-10 Passport 12m/Passport 17m Operator's Manual

Alarm Limit

SVRI Low

SV High

SV Low

SVI High

Range

0 to(high limit-50)DS·m 2 /cm 5 or 0 to(high limit-5)kPa-s-m 2 /l

(Low limit+5) to 300 ml

0 to (high limit-5)ml

(Low limit+5) to 200 ml /m

2

Step

5 ml

Applicable For

SVI Low

RVEF High

RVEF Low

SVV High

SVV Low

0 to (high limit-5)ml /m 2

(Low limit+5) to 100 %

0 to (High limit-5)%

(Low limit+1) to 100 %

0 to (high limit-1)%

5 ml /m

2

5 %

1%

Vigilance II

®

Monitor

Vigileo

TM

Monitor, and

EV1000 Monitor

16.8

Understanding SvO

2

/ScvO

2

Parameters

When the monitor is connected to the Vigilance II® monitor/VigileoTM monitor/EV1000 monitor, you can view all the oxygenation parameters.

To view the oxygenation parameters, following this procedure:

1. Select the SvO

2

parameter window to enter the [

SvO

2

Setup

] memu, or select the ScvO

2

parameter window to enter the [

ScvO

2

Setup

] memu.

2. Select [

Oxygenation Parameters >>

].

16.8.1

Oxygenation Parameters for Vigilance II® Monitor

Abbreviation Unit Full Spelling Measurement Range

SvO

2

ScvO

2

SaO

2

DO

2

VO

2

O

2

EI

%

% mixed venous oxygen saturation central venous oxygen saturation

% arterial oxygen saturation ml/min oxygen delivery ml/min oxygen consumption

% oxygen extraction index

0 to 99%

0 to 99%

40 to 100%

0 to 2000 ml/min

0 to 999 ml/min

0.0 to 99.9%

16.8.2

Oxygenation Parameters for Vigileo

TM

Monitor

Abbreviation Unit Full Spelling Measurement Range

SvO

2

% mixed venous oxygen saturation 0 to 99%

Resolution

1

1

1

1

1

0.1

Resolution

1

ScvO

2

% central venous oxygen saturation 0 to 99% 1

Passport 12m/Passport 17m Operator's Manual 16-11

DO

2

DO

2

I

VO

2

VO

2

I

16.8.3

Oxygenation Parameters for EV1000 Monitor

Abbreviation Unit Full Spelling

SvO

2

ScvO

2

VO

2 e

VO

2

Ie

Hb

SpO

2

%

% mixed venous oxygen saturation central venous oxygen saturation ml/min oxygen delivery ml/min/m

2 oxygen delivery index

Measurement Range

0 to 99%

0 to 99%

0 to 2000 ml/min

10 to 5000 ml/min/m

2 ml/min oxygen consumption 0 to 999 ml/min ml/min/m 2 oxygen consumption index ml/min estimated oxygen consumption index when ScvO

2

is being monitored

10 to 5000 ml/min/m

0 to 999 ml/min

2 ml/min/m 2 g/L, g/dl or mmol/L

% estimated oxygen consumption index hemoglobin

10 to 999 ml/min/m 2

20.0 to 318.0-g/L

2.0 to 31.8 g/dl

1.2 to 19.7mmol/L arterial oxygen saturation from pulse

0 to 100 % oximetry

1

1

1

1

Resolution

1

1

1

1

0.1

1

16.9

Understanding the SvO

2

/ScvO

2

Display

Depending on the setup of the Vigilance II® monitor, Vigileo TM monitor, or EV1000 monitor, the monitor displays either the SvO

2

numeric area or ScvO

2

numeric area.

SvO

2

value

SQI bar graphbar

ScvO

2

value

SQI bar graphbar

NOTE

When EV1000 monitor works in ClearSight mode, neither the SvO

2

numberic area nor the ScvO

2

numberic area is displayed on the monitor’s screen.

16-12 Passport 12m/Passport 17m Operator's Manual

16.10

Changing SvO

2

/ScvO

2

Settings

16.10.1

Setting Signal Output

This monitor outputs analog signals for the Vigilance II® monitor. You can select [

Signal Output Setup >>

] from the

[

SvO

2

Setup

] menu or [

ScvO

2

Setup

] menu to set the source of MAP signals. You can also select [

Simulated High Value

] or [

Simulated Low Value

] to provide simulated high value or low value signals for the Vigilance II® monitor. Refer to the

Vigilance II® Operator’s Manual for the monitor calibration.

16.10.2

Selecting Alarm Properties

When the monitor is connected to the Vigilance II® monitor or EV1000 monitor, select the SvO

2

or ScvO

2

parameter area to enter the [

SvO

2

Setup

] or [

ScvO

2

Setup

] menu and then select [

Alarm Setup

] to set the alarm properties for relevant parameters.

When the monitor is connected to the Vigileo TM monitor, select the SvO

2

or ScvO

2

parameter area to enter the [

Alarm

Setup

] menu and set the alarm properties for relevant parameters.

The following table lists the alarm limits for oxygenation parameters that come from Vigilance II

®

Monitor, Vigileo TM

Monitor, and EV1000 Monitor.

Alarm Limit Measurement Range Resolution Applicable For

SvO

2

High

SvO

2

Low

(Low limit+1) to 99%

0 to (High limit-1)%

ScvO

2

High (Low limit+1) to 99%

1

Vigilance II

®

Monitor, Vigileo TM Monitor, and

EV1000 Monitor

ScvO

2

Low 0 to (High limit-1)%

Passport 12m/Passport 17m Operator's Manual 16-13

FOR YOUR NOTES

16-14 Passport 12m/Passport 17m Operator's Manual

17

Monitoring ScvO

2

17.1

Introduction

Central venous oxygen saturation (ScvO

2

) is measured across spectrophotometry. Spectrophotometry involves the use of light emitting diodes (LED) that produce light of various wavelengths in red and infrared spectra. The light is transmitted to the blood through a fiberoptic in the probe, reflected off the red blood cells and transmitted back through a separate fiberoptic to an optical module. The central venous oxygen saturation is calculated through the analysis of the reflected spectra.

Setup key

Calibration key

CeVOX cable connector

17.2

Safety Information

WARNING

ScvO

2 monitoring is limited to adult and pediatric patients.

NOTE

For ScvO

2 monitoring using the Edwards Vigilance II monitor, Vigileo TM monitor, or EV1000 monitor, please see

Chapter 16 Monitoring CCO/ SvO

2

.

Passport 12m/Passport 17m Operator's Manual 17-1

17.3

Performing ScvO

2

Measurements

1. CeVOX optical module

To perform the ScvO

2

measurements:

2. CeVOX fiberoptic probe 3. Central venous catheter

1. Apply the central venous catheter.

2. Place one end of the fiberoptic probe into the central venous catheter through the distal lumina, and connect the other end to the CeVOX optical module. Then plug the CeVOX cable into the ScvO

2 module.

3. If you see the message [

Calibration Required

], calibrate the ScvO

2

before performing the measurements. For detailed information on ScvO

2

calibration, please see section

17.4 ScvO2 Calibration.

4. Check the reading in the ScvO

2

parameter window.

WARNING

To avoid installation failure, ensure that the proper fiberoptic probe is selected.

Incorrect placement of the fiberoptic probe can lead to vessel perforation. Therefore check the correct position of the probe as indicated in the probe’s instructions for use.

17-2 Passport 12m/Passport 17m Operator's Manual

17.4

ScvO

2

Calibration

Regular in vivo calibration is required using blood gas analysis of a central venous blood sample to ensure accurate measurement of continuous ScvO

2.

For optimal accuracy, it is recommended that an in vivo calibration be performed at least every 24 hours or if hemoglobin is changing (for more details, check the notes below).

To perform calibration:

1. Check the central venous catheter and CeVOX probe for proper placement.

2. Check the quality of the signal. The Signal Quality Indicator (SQI) is used for assessing the quality of fiberoptical signals during probe placement, calibration and measurement. The signal quality is indicated by bars of different height levels. Generally, the higher the level, the better the signal.

3. Withdraw a sufficient amount of central venous blood from the side port of the CeVOX probe to avoid intermixture of infusion/injection with the withdrawn blood.

4. Slowly withdraw 2ml blood from the side port of the CeVOX probe. Do not pull too strongly in order to avoid a hemolysis.

5. Immediately confirm by pressing the [

Sample drawn

] button.

6. If necessary put blood sample on ice and perform an analysis by a blood gas analysis device or a laboratory oximeter.

7. Input lab values for Hb/Hct and ScvO

2

and press [

Calibrate

] to confirm.

NOTE

The SQI signal can be affected by the presence of electrosurgical units. Keep electrocautery equipment and cables away from the monitor and use separate power socket if possible.

To achieve optimal accuracy, it is recommended that the entered hemoglobin and hematocrit values are updated when there is a change of 6 % or more in hematocrit, or of 1.8 g/dl (1.1 mmol/l) or more in hemoglobin. A change in hemoglobin may also affect SQI.

Dye (e.g. Indocyanine Green) or other substances, containing dyes which usually modify the light absorption capacities, can lead to faulty measurement values of the oxygen saturation.

17.5

Understanding the ScvO

2

Display

SQI bar graph

ScvO

2

value

Passport 12m/Passport 17m Operator's Manual 17-3

17.6

Understanding ScvO

2

Parameters

Apart from ScvO

2

, the monitor can also monitor DO

2

, VO

2

, DO

2

I, and VO

2

I. You can access the [

ScvO

2

Calibration

] menu from the [

ScvO

2

Setup

] menu and input a SaO

2

value in [

SaO

2

] edit box. The monitor will calculate the values for oxygention parameters automatically, and display these parameters at [

Oxygention Parameters

] in the [

ScvO

2

Setup

] menu. If a parameter value exceeds its normal range, the system will add a “

” or “

” to the right of the parameter.

WARNING

The monitor may only be regarded as a device providing early warning. If there is an indication of a trend towards de-oxygenation of the patient, blood samples must be taken and tested on a laboratory oximeter in order to arrive at a decision concerning the condition of the patient.

17.7

Changing ScvO

2

Settings

17.7.1

Selecting Hb/Hct

To select Hb/Hct:

1. Open the [

ScvO

2

Setup

] menu.

2. Select [

Hb/Hct

] and then select [

Hb

] or [

Hct

].

17.7.2

Selecting Alarm Properties

Select [

Alarm Setup >>

] from the [

ScvO

2

Setup

] menu to set the alarm properties for the relevant parameters.

17-4 Passport 12m/Passport 17m Operator's Manual

18

Monitoring CO

2

18.1

Introduction

CO

2 monitoring is a continuous, non-invasive technique for determining the concentration of CO

2

in the patient’s airway by measuring the absorption of infrared (IR) light of specific wavelengths. CO

2 has its own absorption characteristic and the amount of light passing the gas probe depends on the concentration of the measured CO

2.

When a specific band of

IR light passes through respiratory gas samples, some of IR light will be absorbed by the CO

2

molecules. The amount of

IR light transmitted after it has been passed through the respiratory gas sample is measured with a photodetector. From the amount of IR light measured, the concentration of CO

2

is calculated.

Sidestream/Microstream measurement samples expired patient gas at a constant sample flow from the patient’s airway and analyzes it with a CO

2

sensor built into the CO

2

module.

The sidestream and microstream CO

2

measurement can be used, with specified accessories, with intubated and non-intubated adult, pediatric, and neonatal patients. With intubated patients, a sample of the respiratory gas is drawn from the patient's breathing circuit through an airway adapter and a gas sampling line. With non-intubated patients, the gas sample is drawn through a nasal cannula.

The measurement provides:

A CO

2

waveform

End tidal CO

2

value (EtCO

2

): the CO

2

value measured at the end of the expiration phase.

Fraction of inspired CO

2

(FiCO

2

): the smallest CO

2

value measured during inspiration.

Airway respiration rate (awRR): the number of breaths per minute, calculated from the CO

2

waveform.

Passport 12m/Passport 17m Operator's Manual 18-1

18.2

Identifying CO

2

Modules

From left to right are sidestream CO

2 module (2 slots), sidestream CO

2 microstream CO

2 module

.

1.0 module(1 slot)/sidestream CO

2

2.0 module (1 slot) , an d

1

1

2

2

1

2

3

3

4

5

3

1.

Setup key to enter the CO

2

setup menu

2.

Measure/standby

3.

Gas outlet

4.

CO

2

watertrap seat

5.

Sampling line connector

If you measure CO

2

using the AG module, see the chapter

19 Monitoring AG.

NOTE

The 1-slot sidestream CO

2

2.0 module is supported by the software version 05.40.00 and above.

18.3

Preparing to Measure CO

2

WARNING

Check that the alarm limit settings are appropriate before taking measurement.

Eliminate the exhausted gas before performing the measurement.

NOTE

Perform the measurement in a well-ventilated environment.

18-2 Passport 12m/Passport 17m Operator's Manual

18.3.1

Making a Sidestream CO2 Measurement

To make a sidestream CO

2

measurement:

1. Attach the watertrap to the module and then connect the CO

2

components as shown below.

Watertrap fixer

Watertrap

Sampling line

2. Plug the CO

2

module into the monitor.

3. The CO

2

module needs time to warm up to reach operating temperature. The message [

CO

2

Sensor Warmup

] is displayed during warm-up.

4. After warm-up is finished, you can perform CO

2

measurements.

CAUTION

The watertrap collects water drops condensed in the sampling line and therefore prevents them from entering the module. To avoid blocking the airway, empty the watertrap container whenever half full.

Dispose of accumulated fluids in accordance with hospital policy or your local regulations.

The watertrap has a filter preventing bacterium, water and secretions from entering the module. After long-term use, dust or other substances may compromise the performance of the filter or even block the airway. In this case, replace the watertrap. It is recommended that you replace the watertrap once a month, or when the watertrap is found leaky, damaged or contaminated.

Do not apply adult watertrap to the neonate patient. Otherwise, patient injury could result.

NOTE

To extend the lifetime of the watertrap and module, disconnect the watertrap and set the operating mode to the standby mode when CO

2

monitoring is not required.

Passport 12m/Passport 17m Operator's Manual 18-3

18.3.2

Making a Microstream CO

2

Measurement

Connect the sampling line to the module and then connect the CO

2

components as shown below. After warm-up is finished, you can perform CO

2

measurements.

Connector for sampling line

Sampling line

18.4

Changing CO

2

Settings

18.4.1

Accessing CO2 Menus

By selecting the CO

2

parameter window, you can access the [

CO

2

Setup

] menu.

18.4.2

Setting the CO

2

Unit

To set the CO

2

Unit:

1. Select [

Main Menu

] → [

Maintenance >>

] → [

User Maintenance

] → enter the required password → [

Unit Setup >>

].

2. In the Unit Setup menu, select [

CO

2

Unit

] and then select [

mmHg

], [

%

] or [

kPa

].

18-4 Passport 12m/Passport 17m Operator's Manual

18.4.3

Setting Up Gas Compensations

WARNING

Make sure to use the appropriate compensations. Inappropriate compensations may cause inaccurate measurement values and result in misdiagnosis.

For the sidestream CO

2

module:

1. Select [

CO

2

Setup

].

2. According to the actual condition, set the concentration required for the following compensations:

[

O

2

Compen

]

[

N

2

O Compen

]

[

Des Compen

]

For the microstream CO

2

module, gas compensations are not required.

18.4.4

Setting Up Humidity Compensation

Sidestream and microstream CO

2

modules are configured to compensate CO

2

readings for either Body Temperature and

Pressure, Saturated Gas (BTPS), to account for humidity in the patient’s breath, or Ambient Temperature and Pressure,

Dry Gas (ATPD).

1. ATPD:

P co2

(mmHg)

=

CO

2

(vol%)

×

P amb

/100

2. BTPS:

P

CO2

(mmHg)

=

CO

2

(vol%)

×

(P amb

47)/100

Where,

P

CO2

= partial pressure,

vol%

= CO

2

concentration,

P amb

= ambient pressure, and unit is mmHg.

For the sidestream and microstream CO

2

module, you can set the humidity compensation on or off according to the actual condition.

To set the humidity compensation:

1. Select the CO

2

parameter window to access the [

CO

2

Setup

] menu, and then select [

BTPS Compen

].

2. Select either [

On

] for BTPS or [

Off

] for ATPD, depending on which compensation applies.

Passport 12m/Passport 17m Operator's Manual 18-5

18.4.5

Setting the Apnea Alarm Delay

To set the apnea alarm delay:

1. Select the CO

2

parameter window to access the [

CO

2

Setup

] menu.

2. Select [

Apnea Delay

] and then select the appropriate setting.

The monitor will alarm if the patient has stopped breathing for longer than the selected apnea time. The setting of

[

Apnea Delay

] takes effect simultaneously to the Resp, CO

2

, AG, and RM modules.

WARNING

The respiration measurement does not recognize the cause of apneas. It only indicates an alarm if no breath is detected when a pre-adjusted time has elapsed since the last detected breath. Therefore, it cannot be used for diagnostic purposes.

18.4.6

Choosing a Time Interval for Peak-Picking

For microstream CO

2

modules, you can select a time interval for picking the highest CO

2

as the EtCO

2

and the lowest as the FiCO

2

.

To set the time interval:

1. Select the CO

2

parameter window to access the [

CO

2

Setup

] menu.

2. Select [

Max Hold

].

3. Select [

Single Breath

], [

10 s

], [

20 s

] or [

30 s

]:

[

Single Breath

]: EtCO

2

and FiCO

2

are calculated for every breath.

[

10 s

], [

20 s

], or [

30 s

]: EtCO

2

and FiCO

2

are calculated using 10, 20 or 30 seconds of data.

18.4.7

Setting the Flow Rate

For the sidestream CO

2

module, you can change the sampling rate of respiratory gas in the patient’s airway by setting the flow rate.

To set the flow rate:

1. Select the CO

2

parameter window to access the [

CO

2

Setup

] menu.

2. Select an appropriate setting from [

Flow Rate

].

WARNING

Please consider the patient’s actual bearing capability and then select the appropriate flow rate.

18-6 Passport 12m/Passport 17m Operator's Manual

18.4.8

Setting up the CO2 Wave

Select the CO

2

parameter window to access the [

CO

2

Setup

] menu, in which you can:

 Select [

Wave Type

] and then select [

Draw

] or [

Fill

]:

[

Draw

]: The CO

2

wave is displayed as a curved line.

[

Fill

]: The CO

2

wave is displayed as a filled area.

Select [

Sweep

] and then select the appropriate setting. The faster the wave sweeps, the wider the wave is.

Change the size of the CO

2

waveform by adjusting the wave [

Scale

].

18.4.9

Setting RR Source

To set RR source:

1. Select the CO

2

parameter window to access the [

CO

2

Setup

] menu.

2. Select [

RR Source

] and then select a source or [

Auto

] from the dropdown list.

The [

RR Source

] setting takes effect simultaneously to the Resp, CO

2

, AG and RM modules. For details, please refer to

section

9.9 Setting Respiration Rate (RR) Source.

18.4.10

Barometric Pressure Compensation

Both sidestream and microstream CO

2 modules have the function of automatic barometric pressure compensation (the system automatically measures the barometric pressure which the patient monitor is exposed to).

18.4.11

Entering the Standby Mode

The Standby mode of the CO

2

module relates to the Standby mode of the monitor as follows:

 If the monitor enters the Standby mode, the CO

2

module also enters the Standby mode.

 If the monitor exits the Standby mode, the CO

2

module also exits the Standby mode.

If the CO

2

module enters or exits the Standby mode, it does not affect the monitor.

To enter or exit the Standby mode manually:

 select the hardkey on the module, or

 select [

Operating Mode

] in the [

CO

2

Setup

] menu and then select [

Standby

] or [

Measure

].

When you set the sidestream CO

2

module to the Standby mode, the CO

2

gas sample intake pump automatically sets the sample flow rate to zero. When exiting the Standby mode, the CO

2

module continues to work at the previously set sample flow rate.

For the sidestream CO

2

module, you can set the delay time after which, the CO

2

module will enter the Standby mode if no breath is detected.

Passport 12m/Passport 17m Operator's Manual 18-7

For the microstream CO

2

module, you can also set a period of time after which the CO

2

module enters the Standby mode if no breath is detected, or since the CO

2

module was powered on, or since the CO

2

module was switched to the measuring mode, or the automatic Standby time is re-set.

To set the standby time:

1. Select the CO

2

parameter window to access the [

CO

2

Setup

] menu.

2. Select [

Auto Standby

] and then select the appropriate setting.

18.5

Measurement Limitations

Some adverse effects can influence the CO

2

performance.

CAUTION

The following factors may influence the accuracy of measurement:

Leaks or internal venting of sampled gas

Mechanical shock

Cyclic pressure up to 10 kPa (100 cmH

2

O)

Other sources of interference, if any

Measurement accuracy may be affected by the breath rate and I/E ratio as follow:

 etCO

2

is within specification for breath rate ≤ 60 bpm and I/E ratio ≤ 1:1;

 etCO

2

is within specification for breath rate ≤ 30 bpm and I/E ratio ≤ 2:1.

Measurement accuracy is unspecified for breath rate larger than 60 bpm.

18.6

Leakage test

When the sidestream CO

2 modules need maintenance, the monitor displays the message in the CO

2

waveform window:

[

Need maintenance. Enter CO2 setup menu.

]. Then:

1. Select [

Main Menu

] → [

Maintenance >>

] → [

User Maintenance >>

] → enter the required password → [

Module

Maintenance >>

]

[

Maintain CO2

].

2. Perform leakage test according to the prompt messages on the menu.

18-8 Passport 12m/Passport 17m Operator's Manual

18.7

Troubleshooting the Sidestream CO

2

Sampling System

When the sampling system of the sidestream CO

2

module does not work correctly, check to see if the sampling line is kinked. If the sampling line is not kinked, remove it from the watertrap. If the monitor gives a message indicating the sampling system still does not work correctly, it indicates that the watertrap is blocked and it should be replaced with a new one. Otherwise, the sampling line is blocked and should be replaced with a new one.

18.8

Removing Exhaust Gases from the System

WARNING

Anesthetics: When using the Sidestream or Microstream CO

2

measurement on patients who are receiving or have recently received anesthetics, connect the outlet to a scavenging system to avoid exposing medical staff to anesthetics.

To remove the sample gas to a scavenging system, connect an exhaust tube to the gas outlet connector of the module.

18.9

Zeroing the Sensor

Zero calibration eliminates the effect of baseline drift during CO

2 measurement exerted on the readings and therefore maintains the accuracy of the CO

2

measurements.

For sidestream and microstream CO

2

modules, a zero calibration is carried out automatically when necessary. You can also start a manual zero calibration if necessary.

To manually start a zero calibration:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password

[

Maintain

CO2 >>

].

2. Select [

Calibrate CO2 >>

]

[

Start Zero Cal.

].

Disconnecting the patient airway is not required when performing a zero calibration.

WARNING

Do not rely on the readings during zeroing.

Passport 12m/Passport 17m Operator's Manual 18-9

18.10

Calibrating the Sensor

For sidestream or microstream CO

2

modules, a calibration should be performed once every year or when the readings go

far beyond the range. For details, refer to the chapter

34 User Maintenance

.

18.11

Oridion Information

Oridion Patents

This posting serves as notice under 35 U.S.C.

§

287(a) for Covidien patents: http://www.covidien.com/patents.

No Implied License

Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized CO

2

sampling consumables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device and/or CO

2

sampling consumable.

18-10 Passport 12m/Passport 17m Operator's Manual

19

Monitoring AG

19.1

Introduction

The anaesthetic gas (AG) module measures the patient’s anesthetic and respiratory gases by connecting to the airway of intubated patients or collecting the gases with specified accessories. It also incorporates the features of the O

2

module and BIS module. The AG measurement is applicable for adult, pediatric and neonatal patients.

The AG module determines the concentration of certain gases using the infrared (IR) light absorption measurement. The gases that can be measured by the AG module absorb IR light. Each gas has its own absorption characteristic. The gas is transported into a sample cell, and an optical IR filter selects a specific band of IR light to pass through the gas. For multiple gas measurements, there are multiple IR filters. The higher the concentration of gas in a given volume the more

IR light is absorbed. This means that higher concentration of IR absorbing gas cause a lower transmission of IR light. The amount of IR light transmitted after it has been passed through an IR absorbing gas is measured. From the amount of IR light measured, the concentration of gas present can be calculated.

Oxygen does not absorb IR light as other breathing gases and is therefore measured relying on its paramagnetic properties. Inside the O

2

sensor are two nitrogen-filled glass spheres mounted on a strong rare metal taut-band suspension. This assembly is suspended in a symmetrical non-uniform magnetic field. In the presence of paramagnetic oxygen, the glass spheres are pushed further away from the strongest part of the magnetic field. The strength of the torque acting on the suspension is proportional to the oxygen concentration. From the strength of the torque, the concentration of oxygen is calculated.

NOTE

The AG modules are configured with automatic barometric pressure compensation function.

Perform the measurement in a well-ventilated environment.

19.2

Identifying AG Modules

The AG module can identify two anesthetic gases in a mixture automatically and distinguish between them according to their contributions to the MAC value for display as the primary and secondary anesthetic agents. The AG module cannot be used outside healthcare facilities or during patient transfer.

Example of three-slot AG module

Setup key

Measure/standby

Gas outlet

Connector for BIS sensor

AG watertrap receptacle

Passport 12m/Passport 17m Operator's Manual 19-1

For details on BIS, refer to the chapter

20

Monitoring Bispectral

Index (BIS

)

.

Example of two-slot AG module

Setup key

Gas outlet

Measure/standby

Indicator

AG watertrap receptacle

Connector for BIS sensor

NOTE

The

two-slot

AG module

is only available for the U.S.

19.3

Understanding the AG Display

The AG module can send waves and numerics for all measured anesthetic gases for display on the monitor, including:

CO

2

, O

2

, N

2

O and AA waves

 awRR: airway respiratory rate

MAC: minimum alveolar concentration

End tidal (Et) and fraction of inspired (Fi) numerics for CO

2

, O

2

, N

2

O and AA

19-2

Where AA represents Des (desflurane), Iso (isoflurane), Enf (enflurane), Sev (sevoflurane), or Hal (halothane).

The AA waveform area displays the primary anesthetic gas’s waveform. The O

2

waveform will be displayed only when the

O

2

waveform is currently switched on.

Passport 12m/Passport 17m Operator's Manual

WARNING

To avoid explosion hazard, do not use flammable anesthetic agent such as ether and cyclopropane with this equipment.

19.4

MAC Values

Minimum alveolar concentration (MAC) is the minimum concentration of the agent in the alveoli. It is a basic index to indicate the depth of anesthesia. The standard ISO 80601-2-55 defines MAC as this: alveolar concentration of an inhaled anesthetic agent that, in the absence of other anesthetic agents and at equilibrium, prevents 50% of patients from moving in response to a standard surgical stimulus.

Minimum alveolar concentration (MAC) values are listed below:

Agent

1 MAC

Des

6%*

Iso

1.15%

Enf

1.7%

Sev

2.1%

Hal

0.77%

N2O

105%**

* indicates the data is taken from a 25-year-old patient.

** indicates 1 MAC nitrous oxide can only be reached in hyperbaric chamber.

NOTE

The MAC values shown in the table above are those published by the U.S. Food and Drug Administration for a healthy 40-year-old adult male patient.

In actual applications, the MAC value may be affected by age, weight and other factors.

The formula to calculate the MAC value is as follows:

MAC

=

i

N

1

=

0

EtAgent

i

AgentVol

age

i

Where N is the number of all agents (including N

2

O) that the AG module can measure, EtAgenti is the concentration of each agent, and AgentVol age i is the concentration of each agent at 1 MAC with age correction.

The formula for calculating age correction of 1 MAC is:

MAC age

=

MAC

40

×

10

(

0.00269

×

(age

40))

For example, the Des concentration at 1 MAC of a 60-year old patient is

6%

×

10

(

0.00269

×

(60

40))

=

6%

×

0.88

.

The AG module measures there are 4% of Des, 0.5% of Hal and 50% of N

2

O in the patient’s end-tidal gas:

MAC

=

6%

4.0%

×

0.88

+

0.5%

0.77%

×

0.88

+

50%

105%

×

0.88

=

2.04

NOTE

The formula above is only suitable for patients who are older than one year. If the patient is less than one year, the system uses one year to do age correction.

Passport 12m/Passport 17m Operator's Manual 19-3

19.5

Preparing to Measure AG

To prepare for the AG measurement:

1. Select an appropriate watertrap according to patient category and attach it to the module.

2. Connect the gas sample line to the watertrap connector.

3. Connect the other end of the gas sampling line to the patient via the airway adapter.

4. Connect the gas outlet to a scavenging system using an exhaust tube.

Airway adapter

AG module

Gas sample line

Exhaust tube

Connect to the patient

5. Insert the AG module into the SMR or the module rack of the monitor and the monitor will prompt [

AG Startup

].

Then the AG module starts to warmup and at the same time the monitor prompts [

AG Warmup

]. After 45 seconds, the AG module enters the iso accuracy mode. After 10 minutes, the module enters the full accuracy mode.

WARNING

Perform the AG leakage test every time before the AG measurement. For instructions on how to perform the

leakage test, see

34.7AG Leakage Test

.

Make sure that the connections are tight. Any leak in the system can result in erroneous readings due to ambient air mixing with patient gases.

Do not apply adult watertrap to the neonate patient. Otherwise, patient injury could result.

Using high-frequency electrosurgical units may increase the risk of skin burn. In this case, do not use antistatic or conductive respiratory tubing.

Check that the alarm limit settings are appropriate before taking measurement.

19-4 Passport 12m/Passport 17m Operator's Manual

CAUTION

Position the airway adapter so that the part connecting to the gas sample line is pointing upwards. This prevents condensed water from passing into the gas sample line and causing an occlusion.

The watertrap collects water drops condensed in the sampling line and therefore prevents them from entering the module. To avoid blocking the airway, empty the watertrap container whenever half full.

Dispose of accumulated fluids in accordance with the hospital policy or your local regulations.

The watertrap has a filter preventing bacterium, water and secretions from entering the module. After long-term use, dust or other substances may compromise the performance of the filter or even block the airway. In this case, replace the watertrap. Replacing the watertrap once a month is recommended.

19.6

Changing AG Settings

19.6.1

Setting Gas Unit

For N

2

O and AA, the unit of the measured gas is fixed to “%”.

To set the gas unit:

1. Select [

Main Menu

] → [

Maintenance >>

] → [

User Maintenance

] → enter the required password → [

Unit Setup >>

].

2. In the Unit Setup menu, you can select [

CO2 Unit

] or [

O2 Unit

] and then select [

mmHg

], [

%

] or [

kPa

].

19.6.2

Setting the Apnea Alarm Delay

To set the apnea alarm delay:

1. Select the AG parameter area to access the [

AG Setup

] menu.

2. Select [

Apnea Delay

] and select the appropriate setting.

The monitor alarms if the patient has stopped breathing for longer than the preset apnea time. The [

Apnea Delay

] setting is universally applied to the Resp, CO

2

, AG, and RM modules.

WARNING

The respiration measurement does not recognize the cause of apneas. It only indicates an alarm if no breath is detected when a preset time has elapsed since the last detected breath. Therefore, it cannot be used for diagnostic purpose.

Passport 12m/Passport 17m Operator's Manual 19-5

19.6.3

Changing the Sample Flow Rate

In the setup menu for any gas, select [

Flow Rate

] and then choose either:

 [

High

]: 200 ml/min for adult and pediatric patients, and 120 ml/min for neonatal patients.

 [

Med

]: 150 ml/min for adult and pediatric patients, and 90 ml/min for neonatal patients.

 [

Low

]: 120 ml/min for adult and pediatric patients, and 70 ml/min for neonatal patients.

19.6.4

Entering the Standby Mode

For the AG module, the default operating mode is measure. When you set the AG module to the Standby mode, the AG gas sample intake pump automatically sets the sample flow rate to zero. When exiting the Standby mode, the AG module continues to work at the previously set sample flow rate with no need to warm up again. After nearly 1 minute, the module enters the full accuracy mode. The Standby mode of the AG module relates to the Standby mode of the monitor as follows:

If the monitor enters the Standby mode, the AG module also enters the Standby mode.

If the monitor exits the Standby mode, the AG module also exits the Standby mode.

If the AG module enters or exits the Standby mode, it does not affect the monitor.

To enter or exit the Standby mode manually, in the agent’s setup menu, select [

Operating Mode

] and then select

[

Standby

] or [

Measure

]. You can also set a period of time after which the AG module enters the Standby mode automatically if no breath is detected since the last detected breath. To set the Standby time, in the agent’s setup menu, select [

Auto Standby (min)

] and then select the appropriate setting.

19-6 Passport 12m/Passport 17m Operator's Manual

19.6.5

Setting Up the AG Wave

Select the AG parameter area to access the [

AG Setup

] menu, you can:

 Select [

CO

2

Wave Type

] and then select [

Draw

] or [

Fill

]:

[

Draw

]: The CO

2

wave is displayed as a curved line.

[

Fill

]: The CO

2

wave is displayed as a filled area.

Select [

Sweep

] and then select the appropriate setting. The faster the wave sweeps, the wider the wave is.

 Change the size of the waveform by adjusting the scale.

19.6.6

Setting RR Source

To set RR source:

1. Select the AG parameter area to access the [

AG Setup

] menu.

2. Select [

RR Source

] and then select a source or [

Auto

] from the dropdown list.

The [

RR Source

] setting is universally applied to the Resp, CO

2

, AG and RM modules. For details, refer to section

9.9

Setting Respiration Rate (RR) Source.

19.7

Changing the Anesthetic Agent

When the anesthetic agent used on the patient is changed, the AG module detects the mixed anesthetic gas during the transition of two anesthetic agents. The time required for completing the replacement of anesthetic agent depends on anesthesia type (low flow or high flow) and the characteristics of anesthetic agents (pharmacokinetics).

The AG module can identify two anesthetic agents automatically. When the proportion of the primary and secondary anesthetic agents in the mixture changes, the AG module can distinguish between them according to their contributions to the MAC value. Then the primary and secondary anesthetic agents will be exchanged for display.

CAUTION

During the transition of two anesthetic agents, the monitor gives no prompt messages and the MAC value displayed may be inaccurate.

19.8

Measurement Limitations

The following factors may influence the measurement accuracy:

Leaks or internal venting of sampled gas

Mechanical shock

Cyclic pressure up to 10 kPa (100 cmH

2

O)

Other sources of interference, if any

Passport 12m/Passport 17m Operator's Manual 19-7

19.9

Troubleshooting

19.9.1

When the Gas Inlet is Blocked

If the gas inlet (including watertrap, sampling line and airway adapter) is occluded by condensed water, the message

[

AG Airway Occluded

] appears.

To remove the occlusion:

Check the airway adapter for an occlusion and replace if necessary.

Check the sampling line for an occlusion or kinking and replace if necessary.

 Check the watertrap for a build up of water. Empty the watertrap. If the problem persists, replace the watertrap.

19.9.2

When an Internal Occlusion Occurs

Condensed water may enter the module and cause contamination and/or internal occlusions. In this case, the message

[

AG Airway Occluded

] displays.

To remove the occlusion:

Check for airway adapter for an occlusion and replace if necessary.

Check the sampling line for an occlusion or kinking and replace if necessary.

Check the watertrap for a build up of water. Empty the watertrap. If the problem persistes, replace the watertrap.

19.10

Removing Exhaust Gases from the System

WARNING

Anesthetics: When using the AG measurement on patients who are receiving or have recently received anesthetics, connect the outlet to a scavenging system to avoid exposing medical staff to anesthetics.

For the AG module sold in the U.S., you can also connect the outlet to the patient respiration circuit to avoid exposing medical staff to anesthetics.

To remove the sample gas to a scavenging system, connect an exhaust tube to the gas outlet connector of the module.

19-8 Passport 12m/Passport 17m Operator's Manual

20

Monitoring Bispectral Index (BIS)

20.1

Introduction

Bispectral index (BIS) monitoring is for use on adult and pediatric patients within a hospital or medial facility providing patient care to monitor the state of the brain by data acquisition of EEG signals.

The BIS, a processed EEG variable, may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall during general anesthesia or sedation.

The BISx is for single side BIS monitoring, and the BISx4 is for both single side and bilateral BIS monitoring. The BISx4 applies only when the BIS Bilateral Sensor is connected.

The BISx or BISx4 equipment must be used under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use.

Setup key

Check sensor

Indicator

BIS cable connector

Passport 12m/Passport 17m Operator's Manual 20-1

20.2

Safety Information

For patients with neurological disorders, patients taking psychoactive medication, and children below the age of 1 year,

BIS values should be interpreted cautiously.

WARNING

The conductive parts of sensors and connectors should not come into contact with other conductive parts, including earth.

To reduce the hazard of burns in the high-frequency surgical neutral electrode connection, the BIS sensor should not be located between the surgical site and the electro-surgical unit return electrode.

To reduce the hazard of burns during use of brain-stimulating devices (e.g., transcranial electrical motor evoked potential), place stimulating electrodes as far as possible from the BIS sensor and make certain that sensor is placed according to package instructions.

The BIS sensor must not be located between defibrillator pads when a defibrillator is used on a patient connected to the patient monitor.

The BIS component using on our monitor is purchased from Covidien. It is important to recognize this index is derived using solely that company's proprietary technology. Therefore, it is recommended that clinicians have reviewed applicable information on its utility and/or risks in published articles and literature/web site information from Covidien, Inc. or contact that company itself at www.covidien.com if you have clinical-based BIS questions relating to this module portion of the patient monitor. Failure to do so could potentially result in the incorrect administration of anesthetic agents and/or other potential complications of anesthesia or sedation. We recommend that clinicians also review the following practice advisory (that includes a section on BIS monitoring): The American Society of Anesthesiologists, Practice Advisory for

Intraoperative Awareness and Brain Function Monitoring (Anesthesiology 2006;104:847-64). Clinicians are also recommended to maintain current knowledge of FDA or other federal-based regulatory, practice or research information on BIS and related topics.

The BIS is a complex technology, intended for use only as an adjunct to clinical judgment and training.

The clinical utility, risk/benefit and application of the BIS component have not undergone full evaluation in the pediatric population.

20-2 Passport 12m/Passport 17m Operator's Manual

20.3

Understanding the BIS Display

20.3.1

BIS Parameter Area

BIS, the primary parameter, always displays in the parameter area. The other parameters are secondary parameters, and

display only when configured. See section

20.9 Changing the Secondary Parameters

for the configuration of the

secondary parameters.

For single side BIS monitoring, the BIS parameter area displays the following parameters:

1

2

3

4

5

1.

Bispectral Index (BIS)

The BIS numeric reflects the patient’s level of consciousness. It ranges from 100 for wide awake to 0 in the absence of brain activity.

BIS numeric

100

70

60

40

0

Description

The patient is wide awake.

The patient is underdosed but still unlikely to become aware.

The patient is under general anesthesia and loses consciousness.

The patient is overdosed and in deep hypnosis.

The EEG waveform is displayed as a flat line, and the patient has no electrical brain activity.

2.

Signal Quality Index (SQI)

The SQI numeric reflects signal quality and provides information about the reliability of the BIS, SEF, TP, and SR numerics during the last minute. Signal quality is optimal when all five bars of the SQI icon are filled with color. SQI ranges from 0-100%.

1 bar represents SQI in the 1%-20% range.

2 bars represent SQI in the 21%-40% range.

3 bars represent SQI in the 41%-60% range.

4 bars represent SQI in the 61%-80% range.

5 bars represent SQI in the 81%-100% range.

0 to 15%: the numerics cannot be derived.

15% to 50%: the numerics cannot be reliably derived.

50% to 100%: the numerics are reliable.

Passport 12m/Passport 17m Operator's Manual 20-3

20-4

3. Electromyograph (EMG)

EMG bar graph reflects the electrical power of muscle activity and high frequency artifacts. The power range is

30-55 dB. When the EMG indicator is low, it indicates that EMG activity is low. BIS monitoring conditions are optimal when the bar is empty.

1 bar represents power in the 31-35 range.

2 bars represent power in the 36-40 range.

3 bars represent power in the 41-45 range.

4 bars represent power in the 46-50 range.

5 bars represent power greater than 51.

EMG>55 dB: this is an unacceptable EMG.

EMG<55 dB: this is an acceptable EMG.

EMG

30 dB: this is an optimal EMG.

4. Suppression Ratio (SR)

SR numeric is the percentage of time over the last 63-second period during which the EEG is considered to be in a suppressed state.

5. Spectral Edge Frequency (SEF)

The SEF is a frequency below which 95% of the total power is measured.

Other parameters will display if configured:

Total Power (TP)

TP numeric, which only monitors the state of the brain, indicates the power in the frequency band 0.5-30Hz. The useful range is 40-100db.

Burst Count (BC)

A burst means a period (at least 0.5 second) of EEG activity followed and preceded by inactivity. The BC numeric helps you quantify suppression by measuring the number of EEG bursts per minute. This parameter is intended for the BIS module with the Extend Sensor or Bilateral Sensor only. BC numeric is valid only when SQI≥15% and SR≥5%.

For bilateral BIS monitoring, the BIS parameter area displays the following parameters (L: Left brain hemisphere; R: Right brain hemisphere):

Passport 12m/Passport 17m Operator's Manual

 BIS L

 EMG L

 SR L

SEF L

SQI L

TP L

BC L

 sBIS L

BIS R

EMG R

SR R

SEF R

SQI R

TP R

BC R sBIS R sBIS (BIS Variability Index)

This numeric represents the standard deviation of the BIS variable over the last three minutes.

 sEMG L sEMG R sEMG (EMG Variability Index)

This numeric represents the standard deviation of the EMG value over the last three minutes.

 ASYM

Asymmetry (ASYM) is a processed variable indicating the percentage of EEG power present in the left or right hemispheres with respect to total (left and right) EEG power.

Designation ‘L’ of the asymmetry data indicates asymmetry to the left side.

Designation ‘R’ of the asymmetry data indicates asymmetry to the right side.

20.3.2

BIS Waveform Area

The BIS waveform area allows you to view either the EEG waveform or the BIS trend. A secondary parameter’s trend line can also be displayed together with the BIS trend line.

To view the BIS waveform area:

1. Enter the [

BIS Setup

] menu.

2. Select [

Display

] and then select the desired option.

[

EEG

]

[

BIS Trend

]

Passport 12m/Passport 17m Operator's Manual 20-5

The available options for BIS trend superimpose display include: [

BIS+EMG Trend

], [

BIS+SQI Trend

], [

BIS+SR

Trend

], [

BIS L+R Trend

], [

BIS+BC Trend

], [

BIS+sBIS Trend

] or [

BIS+sEMG Trend

], depending on the sensor type.

20.3.3

BIS Expand View

When BIS Bilateral Sensor is used for bilateral monitoring, BIS expand view can be displayed.

To view the BIS expanded view:

1. Enter the [

BIS Setup

] menu.

2. Select [

BIS Expand View >>

].

3. Select [

Display

] and then select [

EEG

], [

BIS Trend

] or [

DSA

].

20.3.3.1

Displaying EEG Waveforms

20-6

You can select the EEG waveforms to be displayed. You can also set the desired scale and sweep speed by selecting corresponding buttons.

Passport 12m/Passport 17m Operator's Manual

20.3.3.2

Displaying BIS Trend

You can select the desired trend lines to be displayed and set the time scale. The artifact mark is displayed at the bottom to indicate SQI value. When SQI<15%, the artifact mark is yellow and the corresponding trend lines of BIS, SR, BC and sBIS are not displayed. When 15%≤SQI<50%, the artifact mark is brown.

20.3.3.3

Displaying Density Spectral Array (DSA)

Maximum

Power

Minimum

Power

The Density Spectral Array (DSA) shows changes in the power spectrum distribution over a certain time period. The DSA represents the power spectra ranging from 49-94 dB. The color bar to the right of the time scale shows the range of colors used to indicate minimum and maximum power. The frequency scale is shown on the horizontal axis with a range from 0-30 Hz.

A white Spectral Edge line is superimposed on the graph where 95% of the total power lies on one side of the line

(toward the inside of the graph) and 5% lies on the other. The Spectral Edge Frequency value (SEF) displays above the graph.

The ASYM graph in the center of the screen shows the degree of asymmetry in EEG power between the left and right hemispheres. The ASYM scale begins at 20% at the center line and runs left or right to 100%. Asymmetry data less than

20% are not displayed on the graph, but are available in the tabular trends.

Passport 12m/Passport 17m Operator's Manual 20-7

20.4

Setting up the BIS Measurement

To set up the BIS measurement:

1. Connect the BISx or BISx4 model to the BIS module.

BIS module

BISx or BISx4 model

Patient cable

BIS attachment clip

BIS sensor

2. Use the attachment clip to secure the BISx or BISx4 model near, but not above the level of the patient’s head.

3. Connect the BISx or BISx4 model to the patient cable.

4. Attach the BIS sensor to the patient following the instructions supplied with the sensor.

NOTE

Make sure the patient’s skin is dry. A wet sensor or a salt bridge could result in erroneous BIS and impedance values.

5. Connect the BIS sensor to the patient interface cable.

CAUTION

Do not use if sensor is dry. To avoid dry out, do not open pack until ready for use. Due to intimate skin contact, reuse may pose risk of infection. If skin rash or other unusual symptom develops, stop using and remove. Limited to short-term use (maximum of 24 hours). Do not cut sensor components, as this can result in improper operation.

20-8 Passport 12m/Passport 17m Operator's Manual

20.5

Auto Impedance Check

Auto impedance checks:

 The combined impedance of the signal electrodes plus the reference electrode. This is done automatically and continuously and does not affect the EEG wave. As long as the impedances are within the valid range, there is no prompt message of this check or its results.

 The impedance of the ground electrode. This is done every ten minutes and takes approximately four seconds. It causes artifact in the EEG wave, and the message [

BIS Ground Checking

] is displayed on the monitor during the check. If the ground electrode does not pass this check, another check is initiated. This continues until the ground electrode passes the check.

By default, auto impedance is switched on. If the auto impedance check interferes with other measurements, it can be switched off.

To switch the function off:

1. Select [

Sensor Check

] in the [

BIS Setup

] menu to open the sensor check window.

2. Set [

Automatic Check

] to [

Off

].

CAUTION

Switching the auto impedance check off will disable automatic prompt to the user of impedance value changes, which may lead to incorrect BIS values. Therefore, this should only be done if the check interferes with or disturbs other measurements.

20.6

Sensor Check

Sensor check measures the exact impedance of each individual electrode. It causes a disturbed EEG wave, and a prompt message is displayed on the monitor.

 The sensor check is automatically initiated when a sensor is connected. To manually start a sensor check, you can either:

Press the hardkey on the BIS module.

Select [

Sensor Check

] in the [

BIS Setup

] menu.

Select [

Start Sensor Check

] in the BIS sensor window.

The sensor check stops automatically if the impedances of all electrodes are within the valid range. To manually stop a sensor check, you can either:

Press the hardkey on the BIS module.

Select [

Stop Sensor Check

] in the sensor check window.

Passport 12m/Passport 17m Operator's Manual 20-9

20.7

BIS Sensor Check Window

To open the sensor check window, select [

Sensor Check

] in the [

BIS Setup

] menu. The graphic in the BIS sensor check window automatically adapts to show the type of sensor you are using, show each electrode as required. Each symbol in the graphic represents an electrode and illustrates the most recently-measured impedance status of the electrodes.

1

2

1

2

1. Sensor Type 2. Expiration Time or Usable Times

Different colors indicate different statuses. The electrode status is displayed below each electrode:

Color Status Description Action

Red

Grey

[

[

Lead Off

Noise

[

High

]

[

Pass]

]

]

The Electrode fell off and has no skin contact.

The EEG signal is too noisy. Impedance cannot be measured.

The impedance is above the limit.

The impedance is within valid range.

Reconnect electrode, or check the sensor-to-skin contact. If necessary, clean and dry skin.

Check the sensor-to-skin contact. If necessary, clean and dry skin.

Yellow

Green No action necessary.

Although BIS may still be measured when the electrode status is [

Noise

] or [

High

], for best performance, all electrodes should be in [

Pass

] status.

20-10 Passport 12m/Passport 17m Operator's Manual

20.8

Choosing the BIS Smoothing Rate

The smoothing rate defines how the monitor averages the BIS value. With the smoothing rate becoming smaller, the monitor provides increased response to changes in the patient’s state. Contrarily, the monitor provides a smoother BIS trend with decreased variability and sensitivity to artifacts.

To change the smoothing rate:

1. Select the BIS parameter window to enter the [

BIS Setup

] menu.

2. Select [

Smoothing Rate

] and then select [

10 s

], [

15 s

] or [

30 s

].

NOTE

When [Smoothing Rate] is set as [10 s] or [30 s], sBIS and sEMG are displayed as invalid values.

20.9

Changing the Secondary Parameters

To choose the desired secondary parameters for display on the screen:

1. Enter the [

BIS Setup

] menu.

2. Select [

Change Secondary Parameter

>>] and then select a maximum of 2 parameters from the pop-up menu.

20.10

Changing the EEG Wave Size

To change the EEG wave size:

1. Enter the [

BIS Setup

] menu.

2. Select [

EEG

] from [

Display

].

3. Select [

Scale

] and then select the appropriate setting.

20.11

Changing the Speed of the EEG Wave

To change the speed of the EEG wave:

1. Enter the [

BIS Setup

] menu.

2. Select [

EEG

] from [

Display

].

3. Select [

Sweep

] and then select the appropriate setting. The faster the wave sweeps, the wider the wave is.

Passport 12m/Passport 17m Operator's Manual 20-11

20.12

Setting the Trend Length

To set the trend length:

1. Enter the [

BIS Setup

] menu.

2. Select a BIS trend option from [

Display

].

3. Select [

Trend Length

] and then select the appropriate BIS time length setting.

20.13

Switching the Filter On or Off

To switch the filter on or off:

1. Enter the [

BIS Setup

] menu.

2. Select [

Filter

] and then select [

On

] or [

Off

]. The default is [

On

].

The filter screens out undesirable interference from the raw EEG wave display. The notch filter includes filters for both 50 and 60 Hz. Filter settings do not affect processing of the trend variables (i.e., BIS, EMG, and SR).

20.14

Covidien Information

Covidien Patents

This posting serves as notice under 35 U.S.C.

§

287(a) for Covidien patents: http://www.covidien.com/patents.

No Implied License

Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized replacement parts or consumables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.

20-12 Passport 12m/Passport 17m Operator's Manual

21

Monitoring NMT (From TOF-Watch® SX Monitor)

21.1

Introduction

This monitor can connect an Organon TOF-Watch® SX monitor for NMT (neuromuscular transmission) monitoring. This monitor can display, store and review measurements from TOF-Watch® SX monitor, as well as present related alarms. On this patient monitor, you can separately set the level of NMT related alarms and switch on or off alarm recording; you can also view TOF-Watch® SX monitor settings of alarm limits and alarm switch.

21.2

Safety

WARNING

TOF-Watch® SX monitor is manufacutred by Organon. This company provides the technology for measuring

NMT parameters. We only provide the connection between this patient monitor and TOF-Watch® SX monitor.

If you have any doubts about the operation and maintenance of the TOF-Watch® SX monitor, please refer to

TOF-Watch® SX monitor operator’s manual or directly contact Organon.

Fully observe TOF-Watch® SX monitor operator’s manual to make settings and to connect the monitor with a patient.

21.3

Connecting a TOF-Watch® SX monitor

The TOF-Watch® SX monitor connects with BeneLink module through an ID adapter. For how to establish connection between the TCM monitor and the BeneLink module, refer to

BeneLink Module Operator’s Manual (PN:

046-011948-00)

.

21.4

NMT Parameters

TOF-Watch® SX monitor provides the following measurements:

TOF-Ratio

 TOF-Count

 PTC

 Single

 Tskin

Passport 12m/Passport 17m Operator's Manual 21-1

21.5

Accessing the NMT Setup menu

You can access the [

+NMT Setup

] menu by selecting the NMT area or selecting

[

Main Menu

]

[

Parameters

>>

]

[

+NMT Setup

>>

]

. In the [

+NMT Setup

]

menu, you can

Toggle [

Alarm Sound

] between [

On

] and [

Off

] to switch on or off NMT alarms on this patient monitor.

View the setup as follows:

Stimulation Current

Stimulation Charge

Pulse Width

TOFs Interval

Transducer Sensitivity

Set alarm level for TOF-Ratio and TOF-Count, switch on or off alarm record.

21.6

NMT Display

8

7

6

1

2

3

1. Parameter unit

2. Alarm status

3. Parameter label

4. Parameter measurement

5. Response amplitude of stimulation

6. Skin temperature

7. Measurement countdown

8. Time of last measurement

4

5

21-2 Passport 12m/Passport 17m Operator's Manual

In the case that you take a measurement in TET50Hz mode, TET100Hz mode, DBS3.3 mode or DBS3.2 mode, only mode label is displayed in the NMT parameter area, which is shown as follows:

Measuring mode

Passport 12m/Passport 17m Operator's Manual 21-3

FOR YOUR NOTES

21-4 Passport 12m/Passport 17m Operator's Manual

22

Monitoring Respiratory Mechanics (RM)

22.1

Introduction

WARNING

RM monitoring is not intended for neonatal patients.

The RM monitoring enables clinicians to understand the ventilator operation and patient respiratory status. In the respiratory mechanics (RM) measurement, the airway pressures are measured, from the part between the patient circuit and intubation tube, using a flow sensor between the Y-piece of patient circuit and the patient connection. The pressure is transferred to the monitor through the tube and measured by a pressure transducer in the RM module. The pressure difference together with the gas concentration information is used to calculate flow. The volume information is obtained by integrating the flow signal. From these three parameters, other parameters such as RR, I:E, Compl, etc. are derived.

Setup key

RM monitoring displays the following waveforms and loops:

 Flow waveform

Paw waveform

 Vol waveform

 FV (flow-volume) loop

 PV (paw-volume) loop

Calibrate

Indicator

RM sensor connector

Passport 12m/Passport 17m Operator's Manual 22-1

RM monitoring provides values for 15 parameters. The 15 parameters can be classified into 4 categories:

1. Paw parameters (not applicable for patients with spontaneous breath)

PIP: peak inspiratory pressure (unit: cmH

2

O)

Pplat:

PEEP:

Pmean:

pressure (unit:cmH

2

O) positive end expiratory pressure (unit: cmH mean pressure (unit: cmH

2

O)

2

O)

2. Flow parameters

PIF: peak inspiratory flow (unit: L/min)

PEF: peak expiratory flow (unit: L/min)

3. Vol parameters

TVi:

TVe:

MVi:

MVe: inspiratory tidal volume (unit: ml) expiratory tidal volume (unit: ml) inspiratory minute volume (L) expiratory minute volume (L)

4. Other parameters

RR: respiratory rate (unit: rpm)

I: E:

Compl: ratio of the inspiratory and expiratory time compliance (unit: ml/cmH

2

O)

FEV1.0:

RSBI: first second forced expiratory volume ratio (unit: %) rapid shallow breathing index (unit: rpm/L)

22.2

Safety Information

WARNING

Check for leaks in the breathing circuit system, as they may significantly affect respiratory mechanics readings.

Match the airway adapter you select to the appropriate patient category. Improper sensor selection may produce excessive ventilation resistance or introduce excessive airway dead space, and inaccurate scales and alarm limits.

Periodically check the flow sensor and tubing for excessive moisture or secretion build-up and purge if necessary.

22-2 Passport 12m/Passport 17m Operator's Manual

NOTE

To avoid the affects of excessive moisture in the measurement circuit, insert the flow sensor airway adapter in the breathing circuit with the tubes upright.

Do not place the airway adapter between the endotracheal tube and an elbow as this may allow patient secretions to block the adapter windows.

Measurement values provided by a ventilator may differ significantly from the values provided by the RM module, due to different locations of the flow sensor.

It is recommended to put a heat moisture exchanger between the ventilator and the flow sensor.

It is recommended to position the connector of flow sensor lower than the pressure sampling line.

Keep the respiration loop away from condensing equipment.

Set the ventilator according to the specified range of the RM module. Or unexpected alarm or inaccurate measurement could result.

22.3

Preparing to Monitor RM

To prepare to monitor RM:

1. Select an appropriate flow sensor in accordance with the patient category.

2. Connect the small tubes of the flow sensor to the RM connector of the module using a color-coded adapter.

3. Insert the flow sensor between the Y-piece of the patient circuit and the patient connection.

Connect to the RM module

Connect to the RM module

Connect to ventilator

Flow sensor

Connect to ventilator

Connect to the patient

Connect to the patient

4. Calibrate the flow sensor according to the procedure described in section

22.8 Calibrating the Flow Sensor.

5. Select [

Sensor Type

] in the [

Calibrate RM

] menu and then choose [

Infant One-time

], [

Disposable

] or [

Reusable

] according to the selected sensor.

22-3 Passport 12m/Passport 17m Operator's Manual

22.4

Understanding the RM Display

The RM display shows either the Paw and Flow waveforms, or the Paw and Vol windows in the waveform area.

1

3

2

2.

5

Flow waveform 3. Paw parameter window

5. Vol parameter window

4

1. Paw waveform

4. Flow parameter window

22.5

Changing RM Settings

22.5.1

Accessing RM Menus

To access the RM menus:

 Select the RM parameter window or waveform area to can access the [

RM Setup

] menu.

Select the Paw wave to access the [

Paw Waveform

] menu.

 Select the Flow wave to access the [

Flow Waveform

] menu.

Select the Vol wave to access the [

Vol Waveform

] menu.

22-4 Passport 12m/Passport 17m Operator's Manual

22.5.2

Setting the Apnea Alarm Delay

To set the apnea alarm delay:

1. Select the RM parameter window or waveform area to access the [

RM Setup

] menu.

2. Select [

Apnea Delay

] and then select the appropriate setting. The monitor will alarm if the patient has stopped breathing for longer than the previously set apnea time.

The [

Apnea Delay

] setting takes effect simultaneously to the Resp, CO

2

, AG, and RM modules.

WARNING

The respiration measurement does not recognize the cause of apneas. It only indicates an alarm if no breath is detected when a pre-adjusted time has elapsed since the last detected breath. Therefore, it cannot be used for diagnostic purposes.

22.5.3

Selecting TV or MV for Display

To select tidal volume (TV) or minute volume (MV) for display in the Vol parameter window:

1. Select the RM parameter window or waveform area to access the [

RM Setup

] menu.

2. Select [

TV/MV

] and then select [

TV

] or [

MV

].

By default, the Vol parameter window displays TV values.

22.5.4

Selecting Flow or Vol Waveform for Display

To select Flow or Vol waveform for display:

1. Select the RM parameter window or waveform area to access the [

RM Setup

] menu.

2. Select [

Flow/Vol

] and then select [

Flow

] or [

Vol

].

22.5.5

Changing the Wave Sweep Speed

To change the sweep speed:

1. Select the RM parameter window or waveform area to access the [

RM Setup

] menu.

2. Select [

Sweep

] and select the appropriate setting. The faster the wave sweeps, the wider the wave is.

Passport 12m/Passport 17m Operator's Manual 22-5

22.5.6

Changing the Wave Scale

To change the wave scale:

1. Select the RM parameter window or waveform area to access the [

RM Setup

] menu.

2. In the [

RM Setup

] menu, select [

Wave Scale >>

].

3. Select the appropriate settings in the pop-up menu.

22.5.7

Setting RR Source

To set the RR (respiration rate) source:

1. Select the RM parameter window or waveform area to access the [

RM Setup

] menu.

2. Select [

RR Source

] and then select a source or [

Auto

] from the dropdown list.

The [

RR Source

] setting takes effect simultaneously to the Resp, CO

2

, AG and RM modules. For details, please refer to

section

9.9 Setting Respiration Rate (RR) Source.

22.6

Understanding the Respiratory Loops

Select [

Respiratory Loop

] in the [

RM Setup

] menu. The following window will be displayed.

22-6

Select to recall a reference loop

Respiratory loop

Respiratory parameters

Passport 12m/Passport 17m Operator's Manual

In this window, you can:

Select [

Save

] to save the respiratory loops in the current respiratory cycle as the reference loops. Up to 4 groups of respiratory loops can be saved, and the saving time is displayed above the respiratory loops.

 Change the respiratory loops displayed on the screen: select [

Setup >>

]→[

Display Loop

] and then select [

PV Loop

] or [

FV Loop

].

Turn on/off reference loop: select [

Setup >>

]→[

Reference Loop

], and then select [

On

] or [

Off

].

Change the size of the PV and FV loops: select [

Setup >>

], and then adjust the [

Paw Scale

], [

Vol Scale

] or [

Flow

Scale

].

 Select parameters for display: select [

Setup >>

]→[

Select RM Parameters >>

], and then select [

All RM Parameters

] or [

Select Desired RM Parameters

]. When you select [

Select Desired RM Parameters

], 6 parameters at maximum can be selected.

Print out all parameters for a reference loop by selecting your desired reference loop and then selecting [

Record

].

22.7

Zeroing the RM Module

A zero calibration is carried out automatically every time the monitor is switched on or the RM module connected, and then a zero calibration will automatically be triggered every 5 minutes. You can also start a manual zero calibration when there is a drift in the zero: in the [

RM Setup

] menu, select [

Zero RM

].

22.8

Calibrating the Flow Sensor

A calibration must be performed every time the RM module is connected to the monitor or the flow sensor is connected.

To calibrate the flow sensor:

1. When calibration is needed, the [

Calibrate RM

] menu pops up automatically. You can also enter the [

Calibrate RM

] menu by selecting [

Calibrate RM >>

] from the [

RM Setup

] menu.

2. Select [

Sensor Type

] and then choose [

Infant One-time

], [

Disposable

] or [

Reusable

] according to the sensor used.

3. Enter the positive and negative factor provided on the flow sensor and select [

Calibrate

].

4. After the calibration is completed successfully, the last calibration time and the message [

Calibration Completed!

] are displayed. Otherwise, the message [

Calibration Failed!

] is displayed.

Passport 12m/Passport 17m Operator's Manual 22-7

FOR YOUR NOTES

22-8 Passport 12m/Passport 17m Operator's Manual

23

Monitoring NMT (from Mindray NMT Module)

23.1

Introduction

The neuromuscular transmission (NMT) module evaluates muscle relaxation of patients under neuromuscular block by measuring the strength of muscle reaction after electrically stimulating the dedicated motor nerve. The electrodes are placed on the patient’s skin over dedicated nerve, a controllable current source delivers stimulation pulses to two skin surface electrodes for the nerve stimulation, and the muscle response is measured with an acceleration sensor.

Calibration key

Start/stop key

Indicator

NMT patient cable connector

NOTE

The

NMT module

is only available for the U.S.

23.2

Safety

WARNING

The NMT measurement is not intended for neonatal patients.

The NMT stimulation should not be applied across or through the head, directly on the eyes, covering the mouth, on the front of the neck, especially the carotid sinus, or from electrodes placed on the chest and the upper back or cross over the heart.

Application of electrodes near the thorax may increase the risk of cardiac fibrillation.

Never apply electrodes to patients in areas where inflammation or injury is evident.

When you are connecting the electrodes or the patient cable, make sure that the connectors do not touch any electrically conductive material including earth.

Patients with nerve damage or other neuromuscular problems may not respond properly to stimulation.

The NMT measurement may show unusual patterns when monitoring muscle paralysis in these patients.

NMT stimulation current pulses may interfere with other sensitive equipment, for example, implanted cardiac pacemakers. Do not use the NMT measurement on patients with implanted medical devices unless so directed by a medical specialist.

Simultaneous use of the NMT with high frequency electrosurgical equipment (ESU) may result in burns at the stimulation site and can also adversely affect measurement accuracy. Make sure the ESU return electrode is properly applied to the patient.

Do not use the NMT in close proximity to shortwave or microtherapy devices, there is a risk of adversely

Passport 12m/Passport 17m Operator's Manual 23-1

WARNING

affecting the NMT measurement.

Never touch the electrodes unless the stimulation has been stopped.

Check each time before use that the material insulating the NMT sensor and the stimulation cable is intact and does not show signs of wear and tear.

Do not use in the presence of flammable anesthetics or gases, such as a flammable anesthetic mixture with air, oxygen or nitrous oxide. Use of the device in such an environment may present an explosion hazard.

CAUTION

NMT monitoring is intended as an adjunct in patient assessment and must be used in conjunction with observation of clinical signs and symptoms.

NMT stimulation can be painful to a non-sedated patient. It is recommended not to stimulate before the patient is adequately sedated.

Pay special attention to current densities exceeding 2 mA r.m.s/cm2 for any electrodes.

It is recommended to use small electrodes in order to obtain a sufficient current density. The diameter of the conductive hydrogel should be 1 inches or less. In order to ensure the quality of the NMT test, be sure only to use stimulation electrodes authorized by FDA.

23.3

Stimulation Modes

The NMT module provides the following stimulation modes. Some stimulation modes require a minimum neurophysiological recovery time and during this recovery phase no new stimulation can be started. So you cannot start a measurement or calibration.

23.3.1

Train-Of-Four (TOF)

TOF mode is recommended for most cases. It is also the factory default setting.

In Train of Four stimulus mode, four stimulation pulses are generated at 0.5 second intervals. Each stimulation of the train causes the muscle to contract. The fade in the individual response to each single stimulation provides a basis for evaluation. The response is measured after each stimulus and the ratio of the fourth to the first response of the TOF sequence is calculated resulting in TOF-Ratio.

When relaxation deepens, the TOF% declines until the fourth response disappears and no TOF% is calculated.

When no TOF% is available, the degree of neuromuscular block is estimated from the number of responses or TOF

Counts. The fewer the response count is detected, the deeper is the relaxation.

If NMT calibration establishes the reference response amplitude, response to the first stimulus (T1) as percentage of the reference value is calculated resulting in T1%.

In TOF mode, the minimum neurophysiological recovery time is 10 seconds. If NMT measurement or calibration is initiated during this period, it will be automatically delayed.

23-2 Passport 12m/Passport 17m Operator's Manual

23.3.2

Single Twitch (ST)

In single twitch (ST) stimulation, the module sends a single electrical pulse and measures the strength of the resulting twitch, the module then calculates the ratio of measured response to the reference twitch resulting in ST-Ratio.

ST mode is practical when using depolarizing relaxants since TOF% does not give any additional information about the patient status. Additionally, when the change of patient’s relaxation level is considered, ST stimulation at a frequency of 1

Hz can indicate the relaxation change in a more real-time way.

23.3.3

Post-Tetanic Count (PTC)

When neuromuscular block deepens, different parameters are needed to measure the response. At first, when the response to the fourth TOF stimulation pulse disappears or the first twitch is very weak, the TOF% is not available and only the number of detected counts can be observed. When stimulation pulses no longer give any stimulation response, you do not get the TOF count either. To monitor the relaxation level, you can start tetanic stimulation and estimate the relaxation level from the Post Tetanic Count (PTC).

PTC stimulation mode starts with a sequence of four current pulses delivered at at2 Hz. If a muscle response is detected, the PTC sequence is stopped and the TOF result is reported. If there is no muscle response, the sequence continues with a five seconds long tetanic stimulation delivered at 50 Hz, followed by a pause of 3 seconds, followed by 20 single current pulses delivered at 1 Hz. The number of detected responses is counted and expressed as PTC. The fewer responses are detected, the deeper is the relaxation.

After tetanic stimulation, NMT measurements and calibration are disabled for 20 seconds and PTC is disabled for 2 minutes.

23.3.4

Double-Burst Stimulation (DBS)

Double Burst Stimulation (DBS) enables better visual observing of the fading in the responses. DBS consists of two separate bursts at an interval of 750 ms, where each burst consists of certain pulses directly after each other at a frequency of 50 Hz. The response ratio of the second to the first burst is calculated resulting in DBS-Ratio, while the number of responses is detected and expressed as DBS Count.

The module supports DBS 3.2 and DBS 3.3. For DBS3.2 mode, the first burst consists of 3 consecutive pulses, and the second burst consists of 2 consecutive pulses. For DBS3.3 mode, both bursts consist of 3 consecutive pulses.

In DBS mode, the minimum neurophysiological recovery time is 15 seconds. If NMT measurement or calibration is initiated during this period, it will be automatically delayed.

Passport 12m/Passport 17m Operator's Manual 23-3

23.4

Preparing for NMT Measurement

To take NMT measurement, connect the NMT patient cable to the NMT module. The following picture shows NMT cable and patient connection.

1. NMT module

3

5

6

2.

3.

NMT patient cable

NMT stimulation cable

1

7

4. NMT sensor cable

5. Proximal electrode

6. Distal electrode

2

4 7. NMT sensor

23.4.1

Skin Preparation

Good electrode-to-skin contact is important for good signal quality. Before applying the electrodes, clean the application site of oil and dirt and avoid placing the electrodes over excessive body hair or lesions. Insufficient cleaning of the skin can cause high skin impedance which could cause the stimulation to stop.

To properly prepare the skin:

1. Select sites with intact skin, without lesion of any kind.

2. Clip or shave hair from application sites as necessary.

3. Thoroughly clean the sites with mild soap and water, leaving no soap residue.

We do not recommend using ether or pure alcohol because this dries the skin and increases the impedance.

4. Dry the skin thoroughly.

WARNING

The NMT measurement is not intended for neonatal patients.

23.4.2

Placing the Electrodes and Sensor

Stimulation of the ulnar nerve in the wrist and acceleration measurements at the adductor pollicis is preferred for routine monitoring.

When monitoring neuromuscular transmission, round surface electrodes with snap connection are a must. Small

(pediatric or neonatal) electrodes are advisable to obtain a sufficient current density. In order to ensure a steady signal quality, be sure only to use CE marked electrodes.

Ensure that the thumb can move freely before applying NMT electrodes and sensor. Follow this procedure to place the electrodes and sensor.

1. Place the distal electrode near the wrist.

2. Place the proximal electrode 2 to 3 cm proximal of the distal electrode.

3. Attach the red cable clamp cable to the proximal electrode.

23-4

4. Attach the black cable clamp cable to the distal electrode.

5. Affix the sensor with its large flat side against the palmar side of the thumb with a piece of tape. The cable should be attached in such a way that it does not 'pull' at the sensor and that movement of the thumb is not obstructed.

Passport 12m/Passport 17m Operator's Manual

The arm used for the NMT measurement should be kept immobile during the whole procedure.

CAUTION

To avoid unintentional electrical shocks always make sure that the NMT stimulation has been stopped before touching the electrodes.

Take care to handle the the NMT sensor, avoiding forcefully striking the sensor.

After repositioning the patient, check that the sensor is still applied and that the thumb can move freely.

NOTE

Correct positioning of the electrodes is important. Small displacements may result in considerable changes in stimulation current requirements. Furthermore, the electrodes must be positioned in such a way to avoid direct stimulation of the muscle.

The electrodes should be applied properly to the patient skin. It has been found that slight pressure on the electrodes may improve the stimulation considerably. Therefore, taping the electrodes to the skin may be advisable.

The more distal the sensor is placed on the thumb, the stronger the acceleration signal. This effect can be used to adjust the signal strength.

23.5

Accessing the NMT Setup Menu

You can access the [

NMT Setup

] menu by selecting the NMT area.

The [

NMT Setup

] menu enables you to perform calibration, and provides quick start to NMT measurements. You can also access the following menu by selecting [

Setup >>

].

Passport 12m/Passport 17m Operator's Manual 23-5

23.6

Calibrating the NMT Measurement

The size of the sensor signal varies from patient to patient. NMT calibration determines supramaximal stimulation current and the reference response amplitude. The reference response amplitude is the twitch at the supramaximal stimulation current when the patient is not paralyzed. The calibration must be done prior to administration of a muscle relaxant drug.

If [

Stimulation Current

] is set to [

Supra (60 mA)

], the module automatically searches for supramaximal current to determine the reference response amplitude. If a value between 0 and 60 mA is selected, the reference response amplitude is determined using the selected stimulation current. For adults, the supramaximal current is usually between

35 and 55 mA.

To starting calibration,

1. Check that settings of [

Stimulation Current

] and [

Pulse Width

] are correct from the [

NMT Setup

] menu.

2. Press the Calibration key on the NMT module, or select [

Calibrate

] from the [

NMT Setup

] menu.

If calibration failed, the NMT module automatically use the default value as the reference amplitude.

NOTE

It is recommended that the patient be anesthetized before setting up the calibration twitch as nerve stimulation can be painful.

Changing the stimulation current or pulse width after calibration invalidates the stored reference data, and therefore recalibration is required.

23.6.1

Starting/Stopping NMT Measurements

To Start NMT measurements,

Press the Start/stop key on the NMT module, or

Select the [

Start NMT

] key from the [

NMT Setup

] menu, or

Select the shortcut key of desired stimulation mode, [

ST 0.1HZ

], [

ST 1HZ

], [

TOF

], [

ST

], [

DBS

], or [

PTC

], from the left side of the [

NMT Setup

] menu.

To Stop NMT measurements, press the Start/stop key on the NMT module, or select [

Stop all NMT

] in the [

NMT Setup

] menu. The measurement is interrupted immediately.

If you need to change the NMT settings after startup, stop the measurements, change the settings, and then restart the measurements.

NOTE

Take care when removing the sensor from the patient. Do not pull on the cable.

23-6 Passport 12m/Passport 17m Operator's Manual

23.7

Change NMT Measurement Settings

From the [

NMT Setup

] menu, you can change stimulation related settings.

23.7.1

Changing Stimulation Mode

The module provides four stimulation modes: TOF, ST, DBS, and PTC, see

23.3 Stimulation Modes

for detail.

In the [

NMT Setup

] menu, set [

Stimulation Mode

] to [

TOF

], [

ST

], or [

DBS

]. To perform tetanic stimulation, directly select the [

PTC

] button.

23.7.2

Changing Stimulation Current

Before calibration and monitoring, confirm that the desired stimulus current is selected.

The current is either supramaximal or manually selected between 0 and 60 mA. For adults, the supramaximal current is usually between 35 and 55 mA. Smaller currents may be desirable for children.

23.7.3

Changing Pulse Width

You can increase the pulse width to increase the effect of the stimulation to help finding the supramaximal current.

Changing pulse width after calibration invalidates the stored reference amplitude.

23.7.4

Changing Measurement Interval

Measurement interval is the time interval between NMT measurements.

This function is not available in the PTC mode.

23.8

Enabling Block Recovery Note

The block recovery note alerts you when the set limit is reached. This indicates that the patient is responding more clearly to the stimuli and the neuromuscular block is decreasing. The note can be used, for example, to help maintain a certain relaxation level.

To enable the note and set the limit for activate the note, select [

Block Recovery

] and set the limit. If [

Off

] is selected, the monitor will not give a note.

23.9

Adjusting Stimulation Tone Volume

You can adjust the volume of NMT stimulation tone by setting [

Stimulation Beep Vol

] from the [

NMT Setup

] menu. The monitor gives a beep at the selected volume at each stimulation pulse if the setting is not [

0

].

Passport 12m/Passport 17m Operator's Manual 23-7

23.10

Understanding NMT Display

Dependent on the selected stimulation mode, the following parameters are provided:

Stimulation mode Parameter label

TOF TOF-Ratio

ST

TOF-Count

ST-Ratio

ST-Count

Unit

%

/

%

/

Maximum bars

4

4

1

1

PTC

DBS

PTC

DBS-Ratio

DBS-Count

/

%

/

The follow picture is an example of NMT display of TOF mode:

/

2

2

10 9

8

1

2

7

1. Parameter unit

2. Parameter value

3. Parameter label

3 4

5 6

4. Stimulation Current

5. Scale: indicates the amplitude of response to stimulation. Bar graph is not shown if calibration is not completed successfully.

6. T1%: response to first stimulus as percentage of the reference amplitude in TOF mode. This value is not shown if calibration is not completed successfully.

7. Bar graph: amplitude of response to the stimulation. The maximum height of the bar graphs displayed is 120%.

8. Measurement interval: The monitor displays “Manual” here if [

Interval

] is set to [

Manual

].

9. Measurement countdown: time to the next measurement. The measurement countdown is not shown if [

Interval

] is set to [

Manual

].

10. Time of last measurement

NOTE

The NMT parameter values darken 15 minutes after the NMT measurement is taken.

The PTC value is shown on the display for 20 seconds after which the NMT module returns to the preset stimulation mode.

23-8 Passport 12m/Passport 17m Operator's Manual

23.11

Recalling Calibration Information

In the situation that the NMT module is power down, or you want move the NMT module to another monitor along with the patient and you want to continue with the already determined calibration information, including stimulation current, pulse width, and reference response amplitude, you can use the recall function.

To recall the calibration information, select the [

Restore Calibration Information

] button from the [

NMT Setup

] menu.

Passport 12m/Passport 17m Operator's Manual 23-9

FOR YOUR NOTES

23-10 Passport 12m/Passport 17m Operator's Manual

24

Device Integration

24.1

Introduction

The BeneLink module is intended for connecting external devices to the Passport17m or Passport 12m patient monitor. It allows the information (patient data, alarms, etc.) from the external device to be displayed, saved, recorded, printed, or calculated through the patient monitor. If the patient monitor is connected with the CMS or eGateway, information from the external device can also be transmitted to the CMS or eGateway.

24.2

Safety Information

WARNING

Devices of the same category cannot be connected to the BeneLink module simultaneously.

A patient monitor supports only one BeneLink module.

The signal labels used on the patient monitor may be different from those given on the external device. For details please see the description of parameters and alarms in corresponding sections of this chapter.

The alarms from the external device may be advanced or delayed before transmission to the patient monitor.

There can be differences between the alarm priorities displayed on your patient monitors and the priorities displayed on the external devices interfaced through BeneLink. Please see the list of Output Signals corresponding with each external device for the alarm priorities used by your patient monitor.

NOTE

Only monitors using software version 05.38.00 or later support BeneLink module software version 2.7 or later.

Passport 12m/Passport 17m Operator's Manual 24-1

24.3

Supported Devices and Device Connection Method

For the list of supported devices, device information, and device connection method, refer to

BeneLink Module

Operator’s Manual (PN: 046-011948-00)

.

24.4

Differences in Displayed Values

In certain cases, there may be differences between the numerics seen on the patient monitor and those seen on the external device. The table below lists some situations and possible reasons.

Situation

Some parameter values are displayed as invalid values on the monitor.

The monitor and the external device may display the parameter values with different numbers of places of decimals.

Possible Reasons

The monitor and the external device may have different parameter configuration or displaying range of values. If the monitor displays a parameter that is not configured in the external device or a parameter value from the external device exceeds the displaying range of the monitor, the corresponding parameter value is displayed on the monitor as an invalid value.

The monitor displays the parameter values from the external device based on its own display rules. Same parameter value is displayed differently when the monitor and external device adopt different numbers of places of decimals of the value for display.

Non-continuously measured values and continuously measured values have the same displaying mode in the monitor.

Non-continuously measured values are displayed on the monitor as latest measured values until a new measurement is performed on the external device.

Differences between the parameter values displayed on the monitor and those displayed on the external device.

Some parameter values are converted to different units during transmission to the monitor so that they can be used for calculations. Sometimes, values from the external device may be advanced or delayed before transmission to the monitor.

NOTE

When the pressure units are converted among cmH

2

O, hPa and mbar, the parameter value remains unchanged, for example, 1cmH

2

O=1hPa=1mbar, which may differ from some external devices.

24-2 Passport 12m/Passport 17m Operator's Manual

24.5

Devices Integrated Window

You can view the external device information in the [

Devices Integrated

] window, which provides the information of both individual devices and multi devices. In the individual device menu, you can select [

Para. Display>>

], [

Units>>

] or

[

Alarms>>

] to set the parameters to be displayed, the parameter units, or view the alarm list.

The parameters in the [

Devices Integrated

] window are displayed in the order of priorities. If the window cannot display all the selected parameters, only parameters with the highest priority are displayed. For the parameters controlled by the external device, their settings display in parentheses.

In the [

Devices Integrated

] window, you can select the [

Multi Devices

] tab to view the parameter information of all the external devices interfaced currently. The displayed parameters are those selected in [

Para. Display

] menu of the individual device window. If the monitor cannot display all the selected parameters, only parameters of the highest priority are displayed.

Passport 12m/Passport 17m Operator's Manual 24-3

24.6

System Functions of the Monitor

24.6.1

Alarms

The monitor does not display the realtime alarms from the external device. However, you can view the current alarm list of the corresponding device by selecting [

Alarms>>

] in the individual device window. The alarm priority is defined by a “*” before each alarm message. An alarm list can display up to 100 alarm messages.

24.6.2

Data Storage

The monitor can save and review the graphic trends, tabular trends, and alarm events of parameters from the external device. In the [

Graphic Trends

] menu and [

Events

] menu, parameters from the external device are displayed in white. In the

[Review

] menu, [

Trend Group

] menu, and [

Print Setup

] menu, a mark “+” is shown before each label of parameters from the external device. Please refer to the parameter list to see which parameters can be saved.

NOTE

Parameters from the external device are saved and displayed according to the time of the patient monitor.

24.6.3

Recording and Printing

Information from the external device can be recorded and printed both in realtime and in graphic and tabular trends by the monitor. In addition, the monitor can also record the history of parameters of the external device.

24-4 Passport 12m/Passport 17m Operator's Manual

25

Monitoring tcGas

25.1

Introduction

This monitor can connect an external device for continuous transcutaneous blood gas monitoring.

This monitor can display, store and review measurements from the external device, as well as present related alarms.

On this monitor, you can separately set the level of tcGas related alarms and switch on or off alarm recording; you can also view the settings of alarm limits and alarm switch at the external device.

This monitor can integrate the following TCM monitors:

TCM CombiM

TCM TOSCA

SenTec Digital Monitor (SDM)

25.2

Safety Information

WARNING

TCM CombiM monitor and TCM TOSCA monitor are manufacutred by Radiometer Medical ApS. This company provides the technology for measuring tcGas parameters. We only provide the connection between this monitor and TCM monitors.

The SenTec Digital Monitor (SDM) is manufacutred by SenTec AG. This company provides the technology for measuring tcGas parameters. We only provide the connection between this patient monitor and the SenTec

Digital Monitor.

If you have any doubts about the operation and maintenance of an external device, please refer to the operator’s manual of the external device or directly contact its manufacturer.

Fully observe the operator’s manual of the external device to make settings and to connect the external device with a patient.

For the intended use and contraindication of the external devices, refer to their operator’s manuals.

Passport 12m/Passport 17m Operator's Manual 25-1

25.3

Connecting an External Device

The TCM monitor connects with BeneLink module through an ID adapter. For how to establish connection between the TCM monitor and the BeneLink module, refer to

BeneLink Module Operator’s Manual (PN: 046-011948-00)

.

25.4

tcGas Parameters

TCM CombiM monitor provides the following measurements:

 tcpCO

2

 tcpO

2

Power

Tsensor

In which, tcpCO

2

and tcpO

2

are primary parameters, and the others are secondary parameters.

TCM TOSCA provides the following measurements:

 tcpCO

2

SpO

2

PR

Power

 Tsensor

In which, tcpCO

2

is primary parameter, and the others are secondary parameters.

SenTec SDMS tcPCO

2

provides the following measurements:

 tcpCO

2

 tcpO

2

SpO

2

 PR

 Power

 Tsensor

In which up to two parameters can be selected as primary parameters and the others are secondary parameters.

Options for primary parameters are tcpCO

2, tcpO

2,

SpO

2, and PR, with tcpCO

2

and tcpO

2 being the defaults.

NOTE

On the SenTec Digital Monitor it is possible to disable/enable the parameters to be monitored. For tcpO2

Monitoring, an OxiVenT™ Sensor with activated PO

2

-option is required. If the SenTec Digital Monitor is operated in neonatal mode, SpO

2

and PR are not supported.

25-2 Passport 12m/Passport 17m Operator's Manual

25.5

Displaying tcGas Parameters

To display tcGas parameters on this monitor:

1. Select the [

Screen Setup

] button to enter the [

Screens

] window.

2. Select [

Screen Setup

]. You can choose where to display the tcGas parameters on the screen.

25.6

Entering the tcGas Setup menu

You can access the [

+tcGas Setup

] menu by selecting the tcGas area or selecting

[

Main Menu

]

[

Parameters >>

]

[

+tcGas Setup>>

] .

In the [

+tcGas Setup

] menu, you can:

Set [

Alarm Sound

] to [

On

] or [

Off

] to switch on or off tcGas alarms on the monitor.

Choose the secondary parameters to be displayed. The tcGas area can display maximum three secondary parameters.

For TCM CombiM monitor, only two secondary parameters, Power and Tsensor, are measured in the [

+tcGas

Setup

] menu. The option [

Change Secondary Parameters >>

] is not available.

Set alarm level for tcGas parameters, and switch on or off alarm record.

25.7

Setting tcpCO

2

/tcpO

2

Unit

To set the tcpCO

2

/tcpO

2

unit:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password

[

Maintain

CO2 >>

].

2. Select [

Unit Setup >>

] and set [

tcpCO

2

/tcpO

2

Unit

] to [

mmHg

] or [

kPa

].

25.8

tcGas Display

If TCM CombiM monitor is connected, the tcGas area is shown as follows:

Passport 12m/Passport 17m Operator's Manual 25-3

If TCM TOSCA monitor is connected, the tcGas area is shown as follows:

If SenTec Digital Monitor is connected, the tcGas area is shown as follows:

25-4 Passport 12m/Passport 17m Operator's Manual

26

Freezing Waveforms

During patient monitoring, the freeze feature allows you to freeze the currently displayed waveforms on the screen so that you can have a close examination of the patient’s status. Additionally, you can select any frozen waveform for recording.

26.1

Freezing Waveforms

To freeze waveforms, select the [

Freeze

] hardkey on the monitor’s front panel.

 The system closes the displayed menu (if any), and opens the [

Freeze

] menu.

All displayed waveforms are frozen, i.e. the waveforms stop being refreshed.

NOTE

The freeze feature does not affect the split-screen view of minitrends, OxyCRG and other patients.

26.2

Viewing Frozen Waveforms

To view the frozen waveforms, you can either:

Select the [

Scroll

] button and then rotate the knob clockwise or counter-clockwise, or

 Directly select the or beside the [

Scroll

] button using a mouse or through the touchscreen.

The frozen waveforms will scroll left or right accordingly. And meanwhile, at the lower right corner of the last waveform, there is an upward arrow. The freeze time displays below the arrow and the initial frozen time is [

0 s

]. With the waveforms scrolling, the freeze time changes at intervals of 1 second. This change will be applied for all waveforms on the screen.

Passport 12m/Passport 17m Operator's Manual 26-1

26.3

Unfreezing Waveforms

To unfreeze the frozen waveforms, you can either:

 Select the button at the upper right corner of the [

Freeze

] menu,

Select the [

Freeze

] hardkey on the monitor’s front panel, or

Perform any other action that causes the screen to be readjusted or opens a menu, such as inserting or removing a module, pressing the [

Menu

] hardkey, etc.

26.4

Recording Frozen Waveforms

To record the frozen waveforms:

1.

In the [

Freeze

] menu, select, in order, [

Wave 1

], [

Wave 2

] or [

Wave 3

] and then select your desired waveforms.

2.

Select the [

Record

] button. The selected waveforms and all numerics at the frozen time are printed out by the recorder.

26-2 Passport 12m/Passport 17m Operator's Manual

27

Review

27.1

Accessing Respective Review Windows

To access the review windows:

1.

Select the [

Review

] QuickKey, or [

Main Menu

]

[

Review >>

].

2.

Select [

Graphic Trends

], [

Tabular Trends

], [

Events

], [

Full Disclosure

] or [

12-lead ECG

] to access their respective review windows.

27.2

Reviewing Graphic Trends

In the [

Review

] menu, select [

Graphic Trends

] to access the following window.

1

2

3

4

5

1.

Event mark area

2.

Time axis

3.

Graphic trends area

4.

Parameter area

5.

Cursor

Events are marked with colors in the event mark area. Red represents high level alarm event. Yellow represents medium/low level alarm event. Green represents manual event.

Passport 12m/Passport 17m Operator's Manual 2 7 -1

In this review window:

Select [

Trend Group

] and you can select a trend group for viewing in the pop-up menu. If [

Custom 1

] or

[

Custom 2

] is selected, you can further select [

Define Trend Group

]. Then you can select the parameters for viewing in the pop-up menu.

Select [

Zoom

] to set the time length of the review window.

Select [

Waves

] to set the number of waves displayed in one page.

To browse the graphic trends, you can either:

Select or graphic trends, or beside [

Scroll

] to move the cursor one step to the left or right to navigate through the

Select or to move the cursor one page to the left or right to navigate through the graphic trends.

A time indicating your current position displays above the parameter area. Numeric measurement values corresponding to the cursor location change as the cursor is moved. The measurement value that triggered a high level alarm has a red background. The one that triggered a medium/low level alarm has a yellow background.

 Select or beside [

Event

] to position the cursor to a different event time.

Select the [

Record

] button to print the currently displayed graphic trends to the recorder.

Select the [

Print

] button to set and print the graphic trends report by the printer. For how to set the graphic trends report, refer to chapter

29 Printing

.

27.3

Reviewing Tabular Trends

In the [

Review

] menu, select [

Tabular Trends

] to access the [

Tabular Trends

] window.

2 7 -2

Events are marked with colors in window’s top area. Red represents high level alarm event. Yellow represents medium/low level alarm event. Green represents manual event.

Passport 12m/Passport 17m Operator's Manual

In this review window:

 Select [

Trend Group

] and you can select a trend group for viewing in the pop-up menu. If [

Custom 1

] or

[

Custom 2

] is selected, you can further select [

Define Trend Group

]. Then you can select the parameters for viewing in the pop-up menu.

 Select [

Interval

] to change the resolution of the trend data and then select the appropriate setting:

[

5 s

] or [

30 s

]: select to view up to 4 hours of tabular trends at 5- or 30-second resolution.

[

1 min

], [

5 min

], [

10 min

], [

15 min

], [

30 min

], [

1 h

], [

2 h

] or [

3 h

]: select to view up to 120 hours of tabular trends at your selected resolution.

[

NIBP

]: select to view the tabular trends when NIBP measurements were acquired.

To browse the tabular trends, you can either:

Select or or beside [

Scroll

] to drag the scrollbar left or right to navigate through the trend database,

Select or to scroll left or right to navigate through the trend database.

The measurement value that triggered a high level alarm has a red background. The one that triggered a medium/low level alarm has a yellow background.

Select or beside [

Event

] to position the cursor to a different event time.

 Select the [

Record

] button to access the [

Record Setup

] menu and set the start and end time of the tabular trends you want to record. This feature is not available when reviewing a patient’s history. By further selecting

[

Record

], you can print the displayed tabular trends to the recorder.

Select the [

Record (All)

] button to print all of the tabular trends to the recorder. The trends will be printed in reverse chronological order from the current time to the last time a patient was admitted to the patient monitor.

Select the [

Print

] button to set and print out the tabular trends report to the printer. For how to set the tabular trends report, refer to chapter

29 Printing

.

27.4

Reviewing Events

27.4.1

Marking Events

During patient monitoring, some events may affect the patient and change the displayed waveforms or numerics displayed on the monitor. To help analyze these waveforms or numeric changes, you can mark these events.

Select [

Main Menu

]

[

Mark Event >>

]. In the pop-up menu, you can select the waves to store after triggering a manual event. You can select [

Trigger Manual Event

] from the [

Mark Event

] menu or the [

Manual Event

] from the

[

Main Menu

] to trigger a manual event and store it at the same time.

When you are reviewing graphic trends, tabular trends or full-disclosure waveforms, the manual event symbol is displayed at the time the event is triggered.

Passport 12m/Passport 17m Operator's Manual 2 7 -3

27.4.2

Reviewing Events

The monitor saves the events in real time. You can review these events.

In the [

Review

] menu, select [

Events

] to access the [

Events

] tab. You can review parameter alarm events, arrhythmia alarm events and manual events. When an event occurs, the system stores all the measurement numerics at the event triggering time and related waveforms 4 seconds, 8 seconds, or 16 seconds before and after the event triggering time, as per the setting of recording length.

NOTE

Pausing or switching off alarms will not be recorded as events. The time of these operations will not be recorded in the system log.

Earlier-recorded events might be overwritten by later ones if the events log reaches capacity.

In the [

Events

] tab:

Select [

Event

] to view the desired events.

Select [

Level

] to view the desired events according to the alarm priority.

After selecting the desired event, you can select [

Details

] to access the following window. In this window, the waveform area displays the waveforms related to the event, and the parameter area displays the parameter values happened at the event trigger time.

NOTE

A total loss of power has no impact on the saved events.

1

1. Waveform area 2. Parameter area

In this window:

You can select

You can select or to navigate through the waveforms. or beside the [

Event

] button to switch between events.

2

2 7 -4 Passport 12m/Passport 17m Operator's Manual

 By selecting the [

Beat Annotation

] button, you can display or hide beat labels above the first ECG waveform on the screen. The white beat labels indiate how the monitor is classifying beats and may help explain suspected, missed, or false arrhythmia calls.

N = Normal

V = Ventricular Ectopic

S = Supraventricular Premature

P = Paced

L = Learning

? = Insufficient information to classify beats

I = Inoperative (for example, Lead Off)

M = No beat detected

When beat labels are displayed above the ECG waveform, they can be printed via the recorder or a laser printer. If you select [

Beat Annotation

] in the [

Events

] tab to display or hide beat labels, these beat labels will also be displayed or hidden in the [

Full Disclosure

] tab, and vise versa.

CAUTION

Ensure that you have selected the optimal leads with the best waveform amplitude and the highest signal-to-noise ratio. Selecting the optimal leads is important for beat detection, beat classification, and

V-Fib detection.

 You can set the desired [

Gain

] for ECG waveform.

 You can set the desired [

Sweep

].

 Select the [

Record

] button to print the displayed alarm events to the recorder.

Select the [

Print

] button to print the displayed alarm events to the pre-connected laser printer.

 Select the [

Events List

] button to view the events list.

Passport 12m/Passport 17m Operator's Manual 2 7 -5

27.5

Reviewing Waveforms

In the [

Review

] menu, select [

Full Disclosure

] to access the following window.

A

B

2 7 -6

A. Waveform area B. Parameter area

In this review window:

 To review full-disclosure waveforms, you need to save waveforms first. Select [

Save Waves >>

] and then select the parameters whose waveforms you want to view. To save full-disclosure waveforms, your monitor must have a

CF storage card.

To view the waveforms, you can either:

Select or the waveforms, or beside the [

Scroll

] button to move the cursor one step left or right to navigate through

Select or to move the cursor one page left or right to navigate through the waveforms.

A time indicating your current position is displayed at the top of the waveform area. Numeric measurement values corresponding to the cursor location are displayed in the parameter area, and change as the cursor is moved.

By selecting the [

Beat Annotation

] button, you can display or hide beat labels above the first ECG waveform on

the screen. For more information on the beat annotation function, refer to the section

27.4.2 Reviewing Events

.

You can change the ECG wave gain by selecting [

Gain

] and then selecting the appropriate setting.

You can change the waveform sweep speed by selecting [

Sweep

] and then selecting the appropriate setting.

You can select the [

Record

] button to print out the first three waveforms and measurement numerics to the recorder.

You can select or beside the [

Event

] button to position the cursor between events.

Passport 12m/Passport 17m Operator's Manual

27.6

Reviewing OxyCRG

In the [

Review

] menu, select [

OxyCRG

] tab to access the following window.

In this window:

Select [

Details

] to view the trends, waveform and measurement numerics of selected parameters.

 Select or beside the [

Scroll

] button to switch between events.

Select or to switch between pages.

Select the button at the lower right corner of this window to change the parameter events to be displayed.

After selecting the [

Details

] button, you can access the following window. In this window, the waveform area displays the trends and waveform of the OxyCRG, and the parameter area displays the parameter values at the event trigger time.

A

A. Waveform area

In this window:

Passport 12m/Passport 17m Operator's Manual

B. Parameter area

B

2 7 -7

Select [

Events List

] to switch to the OxyCRG events list.

 Select [

Setup

] to change the displayed parameters.

 Select or beside the [

Event

] button to adjust the position of the cursor between events.

Select or beside the [

Scroll

] button to move the cursor one step left or right to navigate through the trends and waveform.

Select or to navigate through the parameter trends and waveform.

Select the [

Record

] button to print out the currently displayed trends, waveform, and measurement numerics by the recorder.

 Select the [

Print

] button to print to the independent printer.

2 7 -8 Passport 12m/Passport 17m Operator's Manual

28

Calculations

28.1

Introduction

The calculation feature is available with your monitor. The calculated values, which are not directly measured, are computed based on the values you provide.

You can perform the following calculations:

Dose calculations

Oxygenation calculations

Ventilation calculations

Hemodynamic calculations

Renal calculations

To perform a calculation, select [

Main Menu

]

[

Calc >>

], or the [

Calculations

] QuickKey and then select the calculation you want to perform.

Upon completion of a calculation, calculation results will be automatically sent to the central monitoring system and eGateway.

WARNING

After the calculation is finished, verify the entered values are correct and the calculated values are appropriate. We assume no responsibility for any consequences caused by wrong entries and improper operations.

NOTE

The calculation feature is independent of other monitoring functions and can be therefore used for patients being monitored by other monitors. Any operation in a calculation window does not affect the patient monitoring by the local monitor.

Passport 12m/Passport 17m Operator's Manual 28-1

28.2

Dose Calculations

28.2.1

Performing Calculations

To perform a dose calculation:

1. Select [

Main Menu

]

[

Cal >>

]

[

Dose >>

], or select [

Calculations

] QuickKey

[

Dose >>

].

2. Select [

Patient Cat.

] and [

Drug Name

] and then select the appropriate settings. The dose calculation program has a library of commonly used drugs, of which Drug A through Drug E are for those not specified in this library.

Drug A, B, C, D, E

Aminophylline

Dobutamine

Dopamine

Epinephrine

Heparin

Isuprel

Lidocaine

Nipride

NItroglycerin

Pitocin

3. The system gives a set of default values when the above steps are finished. However, these values cannot be used as the calculated values. The user must enter values following the doctor’s instructions, and only then can the calculated values be used

4. Enter the patient’s weight.

5. Enter other values.

6. Verify if the calculated values are correct.

28.2.2

Selecting the Proper Drug Unit

Each drug has its fixed unit or unit series. Among a unit series, one unit may change to another automatically depending on the entered value.

The units for each drug are as follows:

Drug A, B, C, Aminophylline, Dobutamine, Dopamine, Epinephrine, Isuprel, Lidocaine, Nipride and NItroglycerin use the unit series: g, mg and mcg.

Drug D, Heparin and Pitocin use the unit series: Unit, KU (kilo units) and MU (million units).

Drug E uses the unit: mEq (milli-equivalents).

You must select the proper drug name (A, B, C, D or E) according to the units when you define a drug not listed in this library.

NOTE

For neonate patients, [Drip Rate] and [Drop Size] are disabled.

28-2 Passport 12m/Passport 17m Operator's Manual

28.2.3

Titration Table

To open the titration table, select [

Titration Table >>

] in the [

Dose Calculation

] window after the dose calculation is finished.

In the titration table, when you change:

 [

Reference

]

[

Interval

]

[

Dose Type

]

The titrated values change accordingly.

You can also:

 Select or , or or beside the vertical scrollbar to view more values.

Select [

Record

] to print the displayed titrated values by the recorder.

28.2.4

Drug Calculation Formulas

Abbreviation Unit

Conc.

Dose

Volume g/ml, unit/ml or mEq/ml

Dose/hr, Dose/kg/min ml

Amount

Duration

Drip Rate g, unit, mEq h gtt/min

Formula

Amount / Volume

Rate × Conc.

Rate × Duration

Rate × Duration

Amount/Dose

INF Rate × Drop Size / 60

28.3

Oxygenation Calculations

28.3.1

Performing Calculations

To perform an oxygenation calculation:

1. Select [

Main Menu

]

[

Cal >>

]

[

Oxygenation >>

], or select [

Calculations

] QuickKey

[

Oxygenation >>

].

2. Enter values for calculation.

3. Select the [

Calculate

] button. The system performs a calculation per the current settings and displays the calculated values.

If a calculated value is outside the range, its background will highlight in yellow. You can select [

Range

] to view its normal range in the unit field.

Invalid values are displayed as [

---

].

In the [

Oxygenation Calculation

] window, you can:

Change the pressure unit, Hb unit and oxygen content unit by selecting [

Press. Unit

], [

Hb Unit

] and [

OxyCont

Unit

] and then selecting the appropriate settings. The changes take effect automatically.

Trigger a recording by selecting the [

Record

] button. The displayed oxygenation calculations are printed out by the recorder.

Review the previously performed calculations by selecting [

Review

].

Passport 12m/Passport 17m Operator's Manual 28-3

28.3.2

Entered Parameters

Abbreviation Unit

C.O.

FiO

2

PaO

2

PaCO

2

SaO

2

PvO

2

SvO

2

Hb

CaO

2 mmHg mmHg

% mmHg

% g/L ml/L

CvO

2

VO

2

RQ

ATMP

Height

Weight ml/L ml/min

None mmHg cm kg venous oxygen content oxygen consumption respiratory quotient atmospheric pressure height weight

28.3.3

Calculated Parameters and Formulas

Abbreviation Unit Full spelling

BSA

VO

2

calc

C(a-v)O

2

O

2

ER

DO

2

PAO

2

AaDO

2

CcO

2

Qs/Qt

C.O. calc m 2 ml/min ml/L

% ml/min mmHg mmHg ml/L

%

L/min

L/min

%

Full spelling

cardiac output percentage fraction of inspired oxygen partial pressure of oxygen in the arteries partial pressure of carbon dioxide in the arteries arterial oxygen saturation partial pressure of oxygen in venous blood venous oxygen saturation hemoglobin arterial oxygen content body surface area oxygen consumption arteriovenous oxygen content difference oxygen extraction ratio oxygen transport partial pressure of oxygen in the alveoli alveolar-arterial oxygen difference capillary oxygen content venous admixture calculated cardiac output

Formula

Wt 0.425 × Ht 0.725 × 0.007184

C(a-v)O

2

× C.O.

CaO

2

CvO

2

100×C(a-v)O

2

/ CaO

2

C.O. × CaO

2

FiO

2

/ 100 × (ATMP

47)

PaCO

2

×[ FiO

2

/100

(1

FiO

2

/100) / RQ ]

PAO

2

PaO

2

Hb × 1.34

0.031 × PAO

2

100× [1.34 × Hb × (1

SaO

2

/100)

0.031×

(PAO

2

PaO

2

)] / [1.34 × Hb × (1

SvO

2

/100)

0.031× (PAO

2

PvO

2

)]

VO

2

/ (CaO

2

CvO

2

)

28-4 Passport 12m/Passport 17m Operator's Manual

28.4

Ventilation Calculations

28.4.1

Performing Calculations

To perform a ventilation calculation:

1. Select [

Main Menu

]

[

Cal>>

]

[

Ventilation >>

], or select [

Calculations

] QuickKey

[

Ventilation >>

].

2. Enter values for calculation. If the monitor is connected to an anesthesia machine or a ventilator, the system automatically loads the supported parameter values to the [

Ventilation Calculation

] window.

3. Select the [

Calculate

] button. The system performs a calculation per the current settings and displays the calculated values.

If a calculated value is outside the range, its background will highlight in yellow. You can select [

Range

] to view its normal range in the unit field.

Invalid values are displayed as [

---

].

In the [

Ventilation Calculation

] window, you can:

Change the pressure unit by selecting [

Press. Unit

] and then selecting the appropriate setting. Corresponding pressure values shall convert and update automatically.

Select the [

Record

] button to record the displayed ventilation calculations to the recorder.

Review the previously performed calculations by selecting [

Review

].

28.4.2

Entered Parameters

Abbreviation Unit

FiO

2

RR

% rpm

Full spelling

percentage fraction of inspired oxygen respiration rate

PeCO

2

PaCO

2

PaO

2

TV

RQ

ATMP mmHg mmHg mmHg ml

None mmHg partial pressure of mixed expiratory CO

2 partial pressure of carbon dioxide in the arteries partial pressure of oxygen in the arteries tidal volume respiratory quotient atmospheric pressure

28.4.3

Calculated Parameters and Formulas

Abbreviation Unit Full spelling

PAO

2

AaDO

2

Pa/FiO

2 a/AO

2 mmHg mmHg mmHg

% partial pressure of oxygen in the alveoli alveolar-arterial oxygen difference oxygenation ratio arterial to alveolar oxygen ratio

Formula

(ATMP

47) × FiO

2

/100

PaCO

2

× [FiO

2

/100

(1

FiO

2

/100) / RQ ]

PAO

2

PaO

2

100 × PaO

2

/ FiO

2

100 × PaO

2

/ PAO

2

MV

Vd

Vd/Vt

VA

L/min ml

%

L/min minute volume volume of physiological dead space

(TV × RR) / 1000

TV × (1

PeCO

2

/ PaCO physiologic dead space in percent of tidal volume 100 × Vd/TV alveolar volume (TV

Vd) × RR / 1000

2

)

Passport 12m/Passport 17m Operator's Manual 28-5

28.5

Hemodynamic Calculations

28.5.1

Performing Calculations

To perform a hemodynamic calculation:

1. Select [

Main Menu

]

[

Cal >>

]

[

Hemodynamic >>

], or select [

Calculations

] QuickKey

[

Hemodynamic >>

].

2. Enter values for calculation.

For a monitored patient, [

HR

], [

Art mean

], [

PA mean

] and [

CVP

] are automatically taken from the currently measured values. If you have just performed C.O. measurements, [

C.O.

] is the average of multiple thermodilution measurements. [

Height

] and [

Weight

] are the patient’s height and weight you have entered.

If the monitor does not provide these values, their fields are blank.

For a monitored patient, confirm the entered values.

3. Select the [

Calculate

] button. The system performs a calculation per the current settings and displays the calculated values.

If a calculated value is outside the range, its background will highlight in yellow. You can select [

Range

] to view its normal range in the unit field.

Invalid values are displayed as [

---

].

In the [

Hemodynamic Calculation

] window, you can:

Trigger a recording by selecting the [

Record

] button. The currently displayed hemodynamic calculations are printed out by the recorder.

Review the previously performed calculations by selecting [

Review

].

PAWP

Art Mean

PA Mean

CVP

EDV

Height

Weight

28.5.2

Entered Parameters

Abbreviation Unit

C.O.

HR

L/min bpm mmHg mmHg mmHg mmHg ml cm kg

Full spelling

cardiac output heart rate pulmonary artery wedge pressure artery mean pressure pulmonary artery mean pressure central venous pressure end-diastolic volume height weight

28-6 Passport 12m/Passport 17m Operator's Manual

LCWI

LVSW

LVSWI

RCW

RCWI

RVSW

RVSWI

SV

SI

SVR

SVRI

PVR

PVRI

LCW

EF

28.5.3

Calculated Parameters and Formulas

Abbreviation Unit Full spelling

BSA

C.I. m

2

L/min/m

2 body surface area cardiac index ml ml/m 2 stroke volume stroke index

DS/cm 5 systemic vascular resistance

DS·m

2

/cm

5 systemic vascular resistance index

DS/cm

5 pulmonary vascular resistance

DS·m

2

/cm

5 pulmonary vascular resistance index kg·m left cardiac work kg·m/m 2 g·m g·m/m 2 kg·m kg·m/m

2 g·m g·m/m

2

% left cardiac work index left ventricular stroke work left ventricular stroke work index right cardiac work right cardiac work index right ventricular stroke work right ventricular stroke work index ejection fraction

Formula

Wt

0.425

× Ht

0.725

× 0.007184

C.O. / BSA

C.O. / HR × 1000

SV/ BSA

79.96 × (AP MAP

CVP) / C.O.

SVR × BSA

79.96 × (PAMAP

PAWP) / C.O.

PVR × BSA

0.0136 × APMAP × C.O.

LCW / BSA

0.0136 × APMAP× SV

LVSW / BSA

0.0136 × PAMAP × C.O.

RCW / BSA

0.0136 × PAMAP × SV

RVSW / BSA

100 × SV / EDV

28.6

Renal Calculations

28.6.1

Performing Calculations

To perform a renal calculation:

1. Select [

Main Menu

]

[

Cal >>

]

[

Renal >>

], or select [

Calculations

] QuickKey

[

Renal >>

].

2. Enter values for calculation.

3. Select the [

Calculate

] button. The system performs a calculation per the current settings and displays the calculated values.

If a calculated value is outside the range, its background will highlight in yellow. You can select [

Range

] to view its normal range in the unit field.

Invalid values are displayed as [

---

].

In the [

Renal Calculation

] window, you can:

Trigger a recording by selecting the [

Record

] button. The displayed renal calculations are printed by the recorder.

 Review the previously performed calculations by selecting [

Review

].

Passport 12m/Passport 17m Operator's Manual 28-7

28.6.2

Entered Parameters

Abbreviation Unit

URK

URNa mmol/L mmol/L

Urine

Posm

Uosm

SerNa

Cr

UCr

BUN ml/24h mOsm/ kgH

2

O mOsm/ kgH

2

O mmol/L

µ mol/L

µ mol/L mmol/L

Height

Weight cm kg height weight

28.6.3

Calculated Parameters and Formulas

Abbreviation Unit Full spelling

Full spelling

urine pstassium urinary sodium urine plasm osmolality urine osmolality serum sodium creatinine urine creatinine blood urea nitrogen

URNaEx

URKEx

Na/K

CNa

Clcr

FENa mmol/24h mmol/24h

% ml/24h ml/min

% urine sodium excretion urine potassium excretion sodium potassium ratio clearance of sodium creatinine clearance rate fractional excretion of sodium

Cosm

CH

2

O

U/P osm

BUN/Cr ml/min ml/h

None

None* osmolar clearance free water clearance urine to plasma osmolality ratio blood urea nitrogen creatinine ratio

U/Cr None urine-serum creatinine ratio

*: BUN/Cr is a ratio under the unit of mol.

Formula

Urine × URNa / 1000

Urine × URK / 1000

100 × URNa / URK

URNa × Urine / SerNa

Ucr × Urine / Cr / (BSA / 1.73) / 1440

100 ×

(

URNa × Cr ) / (SerNa × Ucr)

Uosm × Urine / Posm / 1440

Urine × (1

Uosm / Posm) / 24

Uosm / Posm

1000 × BUN / Cr

Ucr / Cr

28.7

Understanding the Review Window

With the review feature, you can review oxygenation, ventilation, hemodynamic and renal calculations. The review window for each calculation is similar. Take the hemodynamic calculations review window for example, you can access it by selecting [

Review

] in the [

Hemodynamic Calculation

] window.

In this review window:

You can select , or to view more values.

 The values that exceed the range are displayed in a yellow background. The [

Unit

] field displays parameter units.

If some parameter values are outside of their normal ranges, you can view their normal range in the [

Unit

] field by selecting [

Range

].

 You can review an individual calculation by selecting its corresponding column and then selecting [

Original

Calc

]. You can record the currently displayed calculations or perform another calculation is this window.

28-8 Passport 12m/Passport 17m Operator's Manual

29

Recording

29.1

Using a Recorder

The thermal recorder prints patient information, measurement numeric, and waveforms, etc.

1 2

4

3

5

1. Start/Stop key: press to start a recording or stop the current recording.

2. Indicator

On: when the recorder works correctly.

Off: when the monitor is switched off.

Flashes: if an error occurred to the recorder, e.g., the recorder runs out of paper.

3. Paper outlet

4. Latch

5. Recorder door

29.2

Overview of Recording Types

By the way recordings are triggered, the recordings can be classified into the following categories:

Manually-triggered realtime recordings.

Timed recordings.

Alarm recordings triggered by an alarm limit violation or an arrhythmia event.

Manually-triggered, task-related recordings.

NOTE

For details about alarm recording, refer to the chapter 7 Alarms .

For details about task-related recordings, refer to respective sections of this manual.

Passport 12m/Passport 17m Operator's Manual 29-1

29.3

Starting and Stopping Recordings

To manually start a recording, you can either:

Press the [

Record

] hardkey on the front of the monitor;

Pressthe hardkey on the front of the recorder module; or

Select the [

Record

] button from the current menu or window.

Automatic recordings will be triggered in the following conditions:

Timed recordings will start automatically at preset intervals.

 If both [

Alarm

] and [

Alm Rec

] for a measurement are set on, an alarm recording will be triggered automatically as alarms occur.

When a recording is in process, to manually stop the recording, you can either:

 Press the [

Record

] hardkey on the front of the monitor;

Press the

hardkey on the front of the recorder module; or

 Select [

Clear All Tasks

] in the [

Record Setup

] menu.

Recordings stop automatically when:

A recording is completed.

The recorder runs out of paper.

When the recorder has an alarm condition.

When a recording is stopped, the following markers will be added:

 Automatically stopped recording: prints two columns of ‘*’ at the end of the report.

 Manually or abnormally stopped recording: prints one column of ‘*’ at the end of the report.

29.4

Setting up the Recorder

29.4.1

Accessing the Record Setup Menu

To access the [

Record Setup

] menu, select [

Main Menu

]

[

Record Setup >>

].

29.4.2

Selecting Waveforms for Recording

The recorder can record up to 3 waveforms at a time. You can select, in order, [

Waveform 1

], [

Waveform 2

] and

[

Waveform 3

] in the [

Record Setup

] menu, and then select the waveforms you want. You can also turn off a waveform recording by selecting [

Off

]. These settings are intended for realtime and scheduled recordings.

29-2 Passport 12m/Passport 17m Operator's Manual

29.4.3

Setting the Realtime Recording Length

After starting a realtime recording, the recording time depends on your monitor’s settings. In the [

Record Setup

] menu, select [

Length

] and then select [

8 s

] or [

Continuous

].

[

8 s

]: record 4-second waveforms respectively before and after current moment.

[

Continuous

]: record the waveforms from the current moment until stopped manually.

29.4.4

Setting the Interval between Timed Recordings

Timed recordings start automatically at preset intervals. Each recording lasts 8 seconds. To set the interval between timed recordings: in the [

Record Setup

] menu, select [

Interval

] and then select the appropriate setting.

29.4.5

Changing the Recording Speed

In the [

Record Setup

] menu, select [

Paper Speed

] and then select [

25 mm/s

] or [

50 mm/s

]. This setting is for all recordings containing waveforms.

29.4.6

Clearing Recording Tasks

In the [

Record Setup

] menu, select [

Clear All Tasks

]. All queued recording tasks are cleared and the current recording is stopped.

29.4.7

Setting the IBP Waveform Overlap Recordings

To switch on or off the recordings for IBP waveform overlapping:

1. Open [

Record Setup

] menu.

2. Set [

IBP Overlap

] to:

[

On

]: If two or more waveforms in the selected waveforms for recording are IBP waveforms, the IBP waveforms will be recorded in the overlapping format.

[

Off

]: IBP waveforms will be recorded normally.

29.5

Loading Paper

To load the paper:

1. Press the latch in the upper right corner of the recorder door to open it.

2. Insert a new roll into the compartment as shown below.

3. Close the recorder door.

4. Check if paper is loaded correctly and the paper end is feeding from the top.

Passport 12m/Passport 17m Operator's Manual 29-3

Paper roll

CAUTION

Use only specified thermal paper. Otherwise, it may cause damage to the recorder’s printhead, the recorder may be unable to print, or poor print quality may result.

Never pull the recorder paper with force when a recording is in process. Otherwise, it may cause damage to the recorder.

Do not leave the recorder door open unless you are reloading paper or removing an issue.

29.6

Removing Paper Jam

If the recorder works incorrectly or produces unusual sounds, check if there is a paper jam first. If a paper jam is detected, follow this procedure to remove it:

1. Open the recorder door.

2. Take out the paper and tear off the draped part.

3. Reload the paper and close the recorder door.

29.7

Cleaning the Recorder Printhead

If the recorder has been used for a long time, deposits of paper debris may collect on the printhead compromising the print quality and shortening the lifetime of the roller. Follow this procedure to clean the printhead:

1. Take measures against the static electricity such as Disposable Wrist Strap for the work.

2. Open the recorder door and take out the paper.

3. Gently wipe around the printhead using cotton swabs dampened with alcohol.

4. After the alcohol has completely dried, reload the paper and close the recorder door.

CAUTION

Do not use anything that may destroy the thermal element.

Do not add unnecessary force to the thermal head.

29-4 Passport 12m/Passport 17m Operator's Manual

30

Printing

30.1

Printer

The monitor can output patient reports via a connected printer. The monitor supports the following printers:

HP LaserJet 1505n

HP LaserJet P2035n

HP LaserJet P4015n

HP LaserJet Pro 400 M401n401dn

HP LaserJet 600 M602

 HP LaserJet Enterprise M605

HP LaserJet Enterprise M608n

The report specifications are:

Paper: A4, Letter

Resolution: 300 dpi

For more details about the printer, see the document accompanying the printer. With product upgrades, the monitor may support more printers without prior notice being given. If you have any questions or doubts about your printer, contact Mindray.

30.2

Connecting a printer

To print the reports or the trend data of a patient, you can choose either print to:

 the local printer

Connect the printer and the monitor directly with a network cable, and then start printing what you want.

 the CMS

If your monitor is connected to a CMS, it is recommended to use the CMS for printing.

Passport 12m/Passport 17m Operator's Manual 29-1

30.3

Setting Up the Printer

To set the printer’s properties, select [

Main Menu

]

[

Print Setup >>

]

[

Printer Setup >>

]. In the [

Printer Setup

] menu, you can:

Select a connected printer.

Select [

Printer

] and then select a connected printer as the monitor’s printer.

Search for a printer

If your selected printer is not in the list or a new printer is added into the network, you can select the [

Search

Printer

] to search for all printers in the network.

Set up the paper

Select [

Paper Size

] and then select [

A4

] or [

Letter

].

30.4

Starting Report Printouts

Reports

ECG reports

Contents

ECG waveforms and relevant parameter values

QT report

Tabular trends

Graphic trends

QT parameter values and waveforms

Depends on the selected parameter group, resolution and time period

Depends on the selected parameter group, resolution and time period

Procedures

Select [

Main Menu

]→[

Print Setup >>

]→[

ECG

Reports >>

]→[

Print

]

Select [

ECG Setup

]

[

QT Analysis >>

]

[

QT View >>

]

[

Print

]

Note: [

QT Analysis

] must be set to [

On

].

Select [

Main Menu

]→[

Print Setup >>

]→[

Tabular Trends

Reports >>

]→[

Print

], or select [

Main

Menu

]→[

Review >>

]→[

Tabular Trends

]→[

Print

]→[

Print

]

Select [

Main Menu

]→[

Print Setup >>

]→[

Graphic Trends

Reports >>]

→[

Print

], or select [

Main

Menu

]→[

Review >>

]→[

Graphic Trends

]→[

Print

]→[

Print

]

Arrh. alarm review

Parameter alarm review

ECG waveforms and relevant parameter values

Depends on the selected alarms

Select [

Print

] in [

Arrh. Events

]

Select [

Main Menu

]→[

Review >>

]→[

Alarms

]→[

Print

]

Realtime waves Depends on the selected waveforms

Select [

Main Menu

]→[

Print Setup >>

]→[

Realtime

Reports >>

]→[

Print

]

29-2 Passport 12m/Passport 17m Operator's Manual

30.5

Stopping Report Printouts

To stop report printouts, select [

Main Menu

]

[

Print Setup >>

]

[

Stop All Reports

].

30.6

Setting Up Reports

30.6.1

Setting Up ECG Reports

You can print ECG reports only from the full-screen, half-screen or 12-lead monitoring screen. To set up ECG reports, select [

Main Menu

]

[

Print Setup >>

]

[

ECG Reports >>

].

[

Amplitude

]: set the amplitude of the ECG waveforms.

[

Sweep

]: set the wave print speed to 25 mm/s or 50 mm/s.

[

Auto Interval

]: if [

Auto Interval

] is set to [

On

], the system will automatically adjust the space between waveforms to avoid overlapping.

[

Gridlines

]: choose whether to show gridlines.

[

12-Lead Format

]: if you select [

12×1

], 12 waveforms will be printed on a paper from top to bottom. If you select

[

6×2

], 12 waveforms will be printed from left to right with 6 waveforms on each half and a rhythm waveform will be printed across the bottom. If you select [

3×4

], 12 waveforms will be printed from left to right with 3 waveforms in each of the 4 columns and a rhythm waveform will be printed across the bottom.

30.6.2

Setting Up Tabular Trends Reports

To set up tabular trends reports, select [

Main Menu

]

[

Print Setup >>

]

[

Tabular Trends Reports >>

].

[

Start time

]: you can set a time period whose trend data will be printed out by setting [

From

] and [

Back

]. For example, if you set [

From

] as 2014-4-2 10:00:00 and [

Back

] as [

2 h

], the outputted data will be from 2014-4-2

08:00:00 to 2014-4-2 10:00:00. In addition, the [

Back

] can be set to either:

[

Auto

]: if [

ReportFormat

] is set to [

Time Oriented

], the report will be printed by time. If [

ReportFormat

] is set to [

Parameter Oriented

], the report will be printed by parameter.

[

All

]: if you select [

All

], all trend data will be printed out. In this case, there is no need to set [

[

Interval

]: choose the resolution of the tabular trends printed on the report.

From

].

[

ReportFormat

]: if you select [

Time Oriented

], the report will be printed by time. If you select [

Parameter

Oriented

], the report will be printed by parameter.

[

Select Parameter >>

]: from the pop-up menu, you can:

[

Currently Displayed Trended Parameters

]: print the parameter trend data selected from the [

Tabular

Trends

].

[

Standard Parameter Group

]: select the standard parameter group for printing.

[

Custom

]: define a parameter group for printing from the parameters displayed in the lower part of the menu.

30.6.3

Setting Up Graphic Trends Reports

To set up graphic trends reports, select [

Main Menu

]

[

Print Setup >>

]

[

Graphic Trends Reports >>

]. Setting up

graphic trends reports is similar with tabular trends reports. Refer to section

30.6.2

Setting Up Tabular Trends Reports for

details.

Passport 12m/Passport 17m Operator's Manual 29-3

30.6.4

Setting Up Realtime Reports

To set up realtime reports, select [

Main Menu

]

[

Print Setup >>

]

[

Realtime Reports >>

].

 [

Sweep

]: set the wave print speed to 12.5 mm/s, 25 mm/s, 50 mm/s, or Auto.

 [

Select Wave >>

]: from the pop-up menu, you can:

[

Current

]: print the currently displayed waves.

[

Select Wave

]: select the desired waves for printing.

30.7

End Case Reports

ECG reports, tabular trends reports, graphic trends reports, NIBP review reports and realtime reports can be set as end case reports. When you discharge a patient, the system will automatically print all contents that are set as end case reports.

For example, to set the ECG report as an end case report:

1. Select [

Main Menu

]

[

Print Setup >>

]

[

ECG Report >>

].

2. Select [

End Case Report

]

[

Set as End Case Report

] and then select [

Ok

] from the pop-up dialog box.

3. Set as described in section

30.6.1 Setting Up ECG Reports

.

30.8

Printer Statuses

30.8.1

Printer Out of Paper

When the printer is out of paper, the printer will not print until the paper is replaced. If there are too many print jobs that are not printed, a printer error may occur. In this case, you need to install paper and then re-send the print request.

Restart the printer if necessary.

To avoid this, ensure that there is enough paper in the printer before sending a print request.

30.8.2

Printer Status Messages

Printer Status Message

Printer unavailable

Possible causes and suggested action

The selected printer is not available. Check if the printer is switched on or correctly connected or installed with paper.

29-4 Passport 12m/Passport 17m Operator's Manual

31

Other Functions

31.1

Analog Output

The monitor provides analog output signals to accessory equipment via the Micro-D connector on the rear of the monitor. To obtain analog output signals, connect the accessory equipment such as an oscillograph, etc. to the monitor and then follow this procedure:

1. Select [

Main Menu

]

[

Analog Output Setup>>

].

2. Select [

Analog Out

.] and then select [

On

].

NOTE

The analog output feature is seldom applied in clinical applications. You can contact your service personnel for more details.

31.2

Exporting the Log

The monitor stores system status information, including failures, abnormity, and technical alarms, into the log. You can export the log to a USB disk.

To export the log,

1. Connect a USB disk to the monitor’s USB connector. See section

2.2.3 Rear View

for the proper location of the USB

connector.

2. Select [

Main Menu

]

[

Maintenance

>>]

[

User Maintenance

>>]

→ enter the required password

[

Others

>>].

3. Select [

Export Log

].

31.3

Transferring Data

You can transfer the patient data saved in the monitor to a PC via a crossover network cable or CF storage card, or within a LAN for data management, for review or print.

31.3.1

Data Export System

You must install the data export system on the intended PC before performing the data transfer operation. Refer to the document accompanying the data export system installation CD-ROM for installation instructions.

The data export system supports patient management, data review, data format conversion, print, etc. in addition to data transfer. Refer to the help file of the data export system software for more details.

Passport 12m/Passport 17m Operator's Manual 31-1

31.3.2

Transferring Data by Different Means

NOTE

Never enter the data transfer mode when the monitor is in normal operation or performs monitoring. You must re-start the monitor to exit the data transfer mode.

Transfer data via a crossover network cable

Before transferring data using a crossover network cable:

1. Connect one end of the crossover network cable to the monitor and the other end to the PC.

2. Set the IP address of the PC. This IP address must be in the same network segment with that of the monitor.

3. Make sure that the data export system is active on the PC.

Then, follow this procedure to transfer data:

1. Select [

Main Menu

]

[

Patient Data >>

]

[

Transfer Data

].

2. Select [

Yes

] from the pop-up message box.

3. Input the IP address already set on the PC.

4. Select [

Start

] to start transferring data.

Transfer data within a LAN

Before transferring data within a LAN:

1. Connect the monitor and the intended PC into the same LAN and acquire the PC’s IP address.

2. Make sure that the data export system is active on the PC.

Follow the same procedure as via a crossover network cable to transfer data.

Transfer data via a CF storage card

1. Power off the monitor and remove the CF storage card from it. Refer to chapter

3 Basic Operations

for details.

2. Run the data export system on the PC.

3. Insert the CF storage card into the card reader that connects the PC.

4. Perform the data transfer operation following the help file of the system software.

31.4

Nurse Call

The monitor provides a nurse call connector to output nurse call signal when a user-defined alarm occurs. To obtain nurse call signal, use the nurse call cable (

PN: 8000-21-10361

) we supply to connect the hospital nurse call system to the nurse call connector of the monitor and then follow this procedure:

31-2 Passport 12m/Passport 17m Operator's Manual

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password.

2. Select [

Others >>

] to access the [

Others

] menu.

3. Select [

Nurse Call Setup >>

] to change the nurse call settings as follows:

Select [

Signal Type

] and then select [

Pulse

] or [

Continuous

].

[

Pulse

]: the nurse call signal is a pulse signal and each pulse lasts one second. When multiple alarms occur simultaneously, only one pulse signal is output. If an alarm occurs but the previous one is not cleared, a new pulse signal will also be output.

[

Continuous

]: the nurse call signal lasts until the alarm ends, i.e. the duration of a nurse call signal is equal to that of the alarm condition.

Select [

Contact Type

] and then select [

Normally Open

] or [

Normally Closed

].

[

Normally Open

]: select if your hospital’s nurse call relay contact is normally open.

[

Normally Closed

]: select if your hospital’s nurse call relay contact is normally closed.

Select [

Alm Lev

] and set the alarm level for nurse call-triggering alarms.

 Select [

Alarm Cat.

] and then select the category to which the nurse call-triggering alarms belong.

Alarm conditions are indicated to nurses only when:

 The nurse call system is enabled,

An alarm that meets your preset requirements occurs, and

The monitor is not in the alarm paused or silence status.

WARNING

To obtain the nurse call signal, use the Mindray approved nurse call cable

(

PN: 8000-21-10361

)

. Otherwise the nurse call function will not work and the monitor may be damaged.

Do not rely exclusively on the nurse call system for alarm notification. Remember that the most reliable alarm notification combines audible and visual alarm indications with the patient’s clinical condition.

NOTE

If no setting is selected from [Alm Lev] or [Alarm Cat.], no nurse call signal will be triggered whatever alarms occur.

31.5

iView System (for Passport 17m only)

The iView system is a subsystem integrated in the monitor to display the PC application program window on the monitor and exchange information.

The iView system of this monitor can be configured with the Windows operating system. You can install and use the required PC application program on the monitor through the Windows operating system.

Passport 12m/Passport 17m Operator's Manual 31-3

31.5.1

Start, Power off and Restart iView System

Start iView System

Select [

Main Menu

]

[

Maintenance>>

]

[

User Maintenance>>

]

→ enter the required password

[

iView

Maintenance>>

]

[

iView Start

]. Then the iView system runs and the ShortCut [

iView

] in the main screen is enabled.

Power off iView system

Select [

Main Menu

]

[

Maintenance>>

]

[

User Maintenance>>

]

→ enter the required password

[

iView

Maintenance>>

]

[

iView Power Off

], and select [

Yes

] in the pop-up. Then the iView system shuts down and the

ShortCut [

iView

] in the main screen is disabled.

Restart iView System

Select [

Main Menu

]

[

Maintenance>>

]

[

User Maintenance>>

]

→ enter the required password

[

iView

Maintenance>>

]

[

iView Restart

], and select [

Yes

] in the pop-up.

NOTE

The Restart, Shutdown, Sleep and Hibernate operations from the [Start] menu of the configured Windows system are ineffective to the iView system. The corresponding operations have to be performed in the

[iView Maintenance>>] menu.

31.5.2

Configuring Application Program ShortCuts

Select [

iView

] and the iView ShortCuts area is displayed. Up to five PC application program ShortCuts can be displayed in this area. You can select from these ShortCuts to use the desired application programme.

To configure the ShortCuts:

1. Select [

Main Menu

]

[

Maintenance>>

]

[

User Maintenance>>

]

→ enter the required password

[

iView

Maintenance>>

]

[

iView Maintenance

].

2. To start the configuration tool, select “Config” on the desktop or select [

Start

]

[

My Computer

] in the lower left corner of the desktop. Run “Config.exe” under the path “C:\Program Files\Mindray”.

31-4 Passport 12m/Passport 17m Operator's Manual

Note

The task bar is hidden automatically and is displayed when the mouse is placed at the bottom of the screen.

3. Select [

Add

] and select the application program to be added from the accessed dialog box. Then select [

Open

] to finish adding the application program.

You can select whether to display ShortCuts. [

Show ShortCut

] is checked by default. If not selected, the application program ShortCuts will not be displayed in the iView ShortCuts area. Not selecting the checkbox usually occurs when the application program is started up indirectly. In this case, add both startup program and started program into [

T8 iView shortcut configuration tool

] and do not select the started program. For example, if you want to start the “iexplore.exe” application program to access“www.mindray.com” through the “IE.bat” batch file, write parameters into the batch file. Then add “IE.bat” and the “iexplore.exe” application programs into the [

T8 iView shortcut configuration tool

] and uncheck the “iexplore.exe” status. Finally, save the setting and exit.

4. Select [

Up

] or [

Down

] to change the display order of ShortCuts.

5. Select the cell under [

ShortCut Name

] to change the name of application program.

6. For the application program that can be started up together with the parameter, select the cell under [

Command

] to configure a parameter of the application program. For example, if you add application program “iexplore.exe” into [

T8 iView shortcut configuration tool

], set [

Command

] to “ www.mindray.com

”. Then in the iView ShortCuts area, select the ShortCut of “iexplore.exe” and the system enters the website “ www.mindray.com

”.

7. Select [

Save&Exit

] to finish ShortCut configuration.

8. Select the checkbox before [

Enable Virtual Keyboard

], and the virtual keyboard can be used after the application program runs.

9. Select [

Never

], [

10

], [

30

] or [

60

] in the drop-down list of [

iView window will be closed after

] to set the time interval for the system to automatically close the iView window. For example, when you set to [

10

], if no operation in the iView window is done in 10 minutes, the iView window will automatically close.

Push the [

Main Menu

] key on the monitor front panel to return to the main screen. iView ShortCuts Area

Passport 12m/Passport 17m Operator's Manual 31-5

31.5.3

Using PC Software

1. Select [

Main Menu

] and select [

iVew

], or select [

iView

] on the main screen directly. The PC software ShortCuts with which your monitor is configured will be displayed.

2. Select the ShortCut corresponding to the PC software you want to use to access the corresponding software screen.

Only one PC software screen can be accessed at a time. iView ShortCuts Area is automatically hidden while the PC software is running. It is automatically displayed when the PC software display is minimized or turned off. You can adjust the size or display position of the software window via mouse.

WARNING

All the waveforms and parameters on the monitor are hidden when the PC software display is maximized.

Pay attention to the risk arising from this operation.

Exit PC software or minimize PC software display when PC software is not in use.

To hide a PC software screen:

Select the button in the upper right corner of the software screen.

 Select another area on the monitor screen.

Push the [

Main Menu

] key or the [

Freeze

] key on the monitor front panel.

If PC software is open, and [

Enable Virtual Keyboard

] is checked in the [

T8 iView shortcut configuration tool

], a virtual keyboard icon hides in the left corner of application window. Select the icon, the virtual keyboard will display.

31.5.4

iView Window Close and Standby

The monitor will automatically close the iView window if you have not done an operation in the iView window for a period of time, depending on the time selected for [

iView window will be closed after

].

To re-access iView window, select the ShortCut [

iView

] in main screen.

When the monitor enters Standby, the iView system will also enter Standby.

31-6 Passport 12m/Passport 17m Operator's Manual

31.5.5

Recover iView System

The USB device for iView maintenance can be used to recover the iView system.

Note

Use the USB device for iView maintenance under the guidance of Mindray authorized personnel. The USB device is only for Passport 17m patient monitor. Never use it on other equipment.

31.5.6

Remote Login

If the monitor is networked via iView network connector, you can remotely log in to the monitor’s built-in Windows system through the PC within the LAN. On the PC, you can view the program running on the monitor and perform remote maintenance. Before exiting remote login, you need to restart the monitor’s Windows system. To restart the monitor’s Windows system, select [

Start

]

[

Run

] on the remote PC and then enter “Shutdown –f –r –t 0”.

31.5.7

Using McAfee Solidifier

McAfee Solidifier is the default installation software on iView Windows system . McAfee Solidifier consolidates the executable files of the system, dynamic link library and batch files by way of dynamic white list. Executable files not included in the white list are held back so as to protect the system. You can update the application program or monitor

Windows system via the McAfee Solidifier.

To update the application program:

1. Enter update status.

Before adding, updating or deleting an application program on iView system, let McAfee Solidifier enter update status first. In this case, select “McAfee Solidifier” on the desktop to enter command line dialog box and then enter command

“sadmin bu”.

Note

Before updating an application program, pay attention to anti-virus measures such as network anti-virus strategy and USB device virus scanning.

2. Enter monitor status.

After adding, updating or deleting an application program of the built-in PC, let McAfee Solidifier enter monitor status.

In this case, select “McAfee Solidifier” on the desktop to enter command line dialog box and then enter command

“sadmin eu”.

Other commonly used commands of McAfee Solidifier include:

 sadmin help: used to view the commonly used commands;

 sadmin status: used to view the status of McAfee.

Passport 12m/Passport 17m Operator's Manual 31-7

31.6

Network Setup

CAUTION

Disconnecting from the network may result in loss of transmitted data, including parameter waveforms and measurements, alarm events, trends and patient data, or cause function failure. In case of network disconnection, check the patient and solve the network problem as soon as possible.

31.6.1

Setting the Network Type

The Passport 12m patient monitor supports both wired and wireless network. The Passport 17m patient monitor supports the wired network only.

To set the network type:

1.

Select [

Main Menu

]

[

Maintenance>>

]

[

User Maintenance>>

]

→ enter the required password

→ select [

Ok

]

.

2.

Select [

Network Setup

>>].

3.

Select [

Monitor Network Setup

>>].

4.

Set [

Network Type

] to [

LAN

] or [

WLAN

].

31.6.2

Wireless Network Connection (For Passport 12m Only)

31.6.2.1

Requirements for Wireless Network

To set up the wireless network, the following requirements should be met:

The Cisco AIR-AP2802I-A-K9, AIR-CAP2602I-A-K9, or AIR-AP1242AG-A-K9, is the recommended wireless Access

Point (AP). Performance cannot be guaranteed if any other wireless APs are used.

Support the 802.11a/b/g/n wireless protocol.

The channel bandwidth should be 20 MHz.

WPA2-PSK security method is recommended for authentication and encryption.

Support WPA-PSK, WPA2-PSK , WPA-Enterprise, and WPA2-Enterprise.

The throughput capacity of each Passport 12m monitor is 2.98 Mbps for the BeneVision CMS. The total throughput of the wireless devices connected to a wireless AP should be less than the effective transmitting capability of the wireless AP. Refer to the AP manufacturer for the AP effective transmitting capability.

Where the monitor is located, the wireless AP signal strength should be no less than -65 dBm.

Where the monitor is located, the signal strength of other Wi-Fi devices of the same channel should be no greater than -85 dBm.

 Where the monitor is located, the signal strength of other Wi-Fi devices of adjacent channels should no greater than -50 dBm.

The recommended distance between the monitor and other non-Wi-Fi wireless devices, including wireless devices at the frequency of 2.4GHz, cellular mobile communication networks, microwave ovens, intercoms, cordless phones and electro-surgical units, is no less than 20 cm.

31-8 Passport 12m/Passport 17m Operator's Manual

CAUTIONS

Always configure the wireless network according to local wireless regulations.

Authentication and encryption other than WPA2-PSK or WPA/WPA2 Enterprise may expose sensitive data and allow a malicious attack on the system.

Keep network authentication information, for example passwords, safe to protect the network from being accessed by unauthorized users.

Disconnecting from the network may result in data loss, including waveforms, numerics, alarm events, trends and patient data, or function failure. Check the patient in case of network disconnection and solve the network problem as soon as possible.

Do not connect non-medical devices to the wireless network.

31.6.2.2

Setting the Wireless Network

To set the wireless network:

1.

Select [

Main Menu

]

[

Maintenance

>>]

[

User Maintenance

>>]

→ enter the required password

→ select [

Ok

].

2.

Select [

Network Setup

>>].

3.

Select [

Monitor Network Setup

>>].

4.

Set [

Network Type

] to [

WLAN

].

5.

Set the desired [

Address Type

].

[

Manual

] : Indicates that the operator will manually enter network settings including IP Address, Subnet Mask,

Gateway.

[

DHCP

]: Indicates that the monitor will automatically acquire network settings from a DHCP server.

Passport 12m/Passport 17m Operator's Manual 31-9

31.6.2.3

Wireless Network Related Problems

Symptoms

Unable to connect the network.

Correction Action

1. Verify that the network is available.

2. Check if the network type is correctly set. For example, if LAN is used, set the network type to LAN and connect the Passport 12m to the network; if WLAN is used, set the network type to WLAN and connect the Passport 12m to the wireless network.

3. Check that the SSID and password of the monitor are consistent with those of the wireless AP.

4. Check for IP address conflicts. If any are found, change the IP addresses to remove the conflict.

5. Check if the Mindray recommended wireless AP is used. If not, verify the AP effective transmission rate meets the throughput requirements of the connected devices.

6. Verify the AP channel bandwidth is 20 MHz.

7. Verify that where the monitor is located, the wireless AP signal strength is no less than -65 dBm.

8. Verify that where the monitor is located, the signal strength of other Wi-Fi devices on the same channel is no greater than -85 dBm.

9. Verify that where the monitor is located, the signal strength of other Wi-Fi devices on adjacent channels is no greater than -50 dBm.

10. Verify that the recommended distance between the monitor and other non-Wi-Fi wireless devices, including wireless devices at the frequency of 2.4GHz, cellular mobile communication networks, microwave ovens, intercoms, cordless phones and electro-surgical units, is no less than 20 cm.

The monitor is frequently off line or disconnects from the network.

The transmission delay is too long.

1. Check if the Mindray recommended wireless AP is used. If not, verify the AP effective transmission rate meets the throughput requirements of the connected devices.

2. Verify the AP channel bandwidth is 20 MHz.

3. Verify that where the monitor is located, the wireless AP signal strength is no less than -65 dBm.

4. Verify that where the monitor is located, the signal strength of other Wi-Fi devices on the same channel is no greater than -85 dBm.

5. Verify that where the monitor is located, the signal strength of other Wi-Fi devices on adjacent channels is no greater than -50 dBm.

6. Verify that the recommended distance between the monitor and other non-Wi-Fi wireless devices, including wireless devices at the frequency of 2.4GHz, cellular mobile communication networks, microwave ovens, intercoms, cordless phones and electro-surgical units, is no less than 20 cm.

7. Verify that no unauthorized devices are connected to the wireless AP.

31.6.3

Viewing the MAC Address

You can get the MAC address from the monitor for network management.

To view the MAC address:

1.

Select [

Main Menu

]

[

Maintenance>>

]

[

User Maintenance>>

]

→ enter the required password

→ select [

Ok

].

2.

Select [

Network Setup

>>].

3.

Select [

Monitor Network Setup

>>].

NOTE

Only the MAC address of the wired NIC is available. The wireless radio MAC address is not available.

31-10 Passport 12m/Passport 17m Operator's Manual

31.6.4

Enabling the Data Encryption

If your enable the data encryption, the patient’s MRN (Medical Record Number), visit number, first name and last name are encrypted when transferring data to the CMS or eGateway.

To enable the data encryption:

1.

Select [

Main Menu

]

[

Maintenance>>

]

[

User Maintenance>>

]

→ enter the required password

→ select [

Ok

].

2.

Select [

Network Setup

>>].

3.

Set [

Network Encrypt Switch

] to [

On

].

31.6.5

Connecting the Monitor to the CMS

The Passport 12m and Passport 17m monitors can be connected to the central station (CMS) through the network. When connected with the CMS,

The monitor transmits waveforms and numerics of parameters (e.g. ECG, Resp, Temp, SpO

2

, CO

2

, NIBP), related alarms and alarm settings, patients' information, operating mode, and and historical data, including trends and events, to the central station.. The waveforms, numerics, alarms, alarm settings, patients' information, operating mode, and historical data displayed on the central station are consistent with the monitor.

When so configured, the central station can transmit alarm settings, patients' information, and operating mode settings to the monitor. The alarm settings, patients' information, and operating mode settings of the monitor are consistent with the central station.

31.6.5.1

Setting the Central Stations for Passport 12m

You can configure up to 30 central stations (CMS) for your monitor. To set the CMSs:

1 Select [

Main Menu

]

[

Maintenance

>>]

[

User Maintenance

>>]

→ enter the required password.

[

Network

Setup

>>]

[

Central Station Setup

>>].

2. Set CMS names and corresponding IP addresses.

Selecting a CMS

If [

Select CM

S] is enabled, you can select the ce ntral station (CMS) for the current monitoring.

To select the CMS, select the prompt message area at the bottom of the screen. Then the selected CMS name will display.

If no CMS is selected, this area displays “???”.

Passport 12m/Passport 17m Operator's Manual 31-11

Clearing the Selected CMS at Startup

You can clear the selected CMS each time the monitor restarts after being powered off for more than 2 minutes.

To clear the selected CMS:

1 Select [

Main Menu

]

[

Maintenance

>>]

[

User Maintenance

>>]

→ enter the required password

[

Others

>>].

2. Set [

Clear CMS IP at startup

] to [

On

].

The selected CMS will not be cleared when only one CMS is configured, or the monitor is restarted within 2 minutes.

This function is switched off by default.

31.6.5.2

Connecting the Passport 17m to the CMS

To connect a Passport 17m monitor to the CMS:

1 Select [

Main Menu

]

[

Maintenance

>>]

[

User Maintenance

>>]

→ enter the required password

[

Network

Setup

>>]

[

Monitor Network Setup

>>].

2 In the [

Monitor Network Setup

] menu, set [

Network Type

] and [

Address Type

].

3 Input the monitor IP address, subnet mask and gateway address if the [

Address Type

] is set to [

Manual

].

4 Admit a Passport 17m monitor on the CMS. Refer to the

H-046-007960-00 BeneVision CMS Ops Manual(FDA)

or

H-046-010879-00 BeneVision CMS Operator's Manual R3

for details of admitting Passport 17m.

31.6.6

Setting the Network Service Quality Level

To set the quality of service (QoS):

1.

Select [

Main Menu

]

[

Maintenance

>>]

[

User Maintenance

>>]

→ enter the required password.

[

Network

Setup

>>]

[

QoS Setup

>>].

2.

Select the desired value for [

Realtime Monitoring

]. This sets the service quality of network connection for important realtime network transactions such as parameter measurements, waveforms, and alarms. The value ranges from 0 to7. The greater the value, the higher priority the network transaction.

3.

Select the desired value for [

Others

]. This sets the service quality of network connection for secondary non-realtime network transactions such as transferring historical data from the monitor to the CMS. The value ranges from 0 to7.

The greater the value, the higher priority the network transaction.

31.6.7

Setting the Multicast Parameters

Multicast parameters must be configured before use on a network.

To set the multicast parameters:

1. Select [

Main Menu

]

[

Maintenance

>>]

[

User Maintenance

>>]

→ enter the required password.

[

Network

Setup

>>]

[

Multicast Setup

>>].

2. Set [

Multicast Addr

] and [

TTL

].

3. Select [

Ok

] to save the setting.

31-12 Passport 12m/Passport 17m Operator's Manual

31.7

MLDAP

MLDAP refers to Mindray LDAP (Lightweight Directory Access Protocol). It is an independent process which can be installed on eGateway or other application server (Windows). MLDAP provides user authorization and authentication.

The MLDAP server is connected to the hospital’s LDAP or ADT server. All monitoring devices are connected to the MLDAP server to perform authorization and authentication for the following operations:

Changing alarm settings

Changing arrhythmia settings

Accessing the [

User Maintenance

] menu

31.7.1

Setting MLDAP

To access the MLDAP server, you should set your monitor as follows:

1. Select [

Main Menu

]

[

Maintenance

>>]

[

User Maintenance

>>]

→ enter the required password.

[

Network

Setup

>>]

[

User Authentication Service Setup

>>].

2. Set IP address and port of the MLDAP server.

31.7.2

Selecting Password for User Authentication

You can select what password is used when changing alarm settings, arrhythmia settings, and accessing the

Maintenance menu. To select the passwords:

1. Select [

Main Menu

]

[

Maintenance

>>]

[

User Maintenance

>>]

→ enter the required password.

[

Password

Protection

>>].

2. Select [

Alarm Setup

] to set password for changing alarm settings.

[

No Password

]: changing alarm settings is not password protected.

[

Local Password

]: changing alarm ON/OFF switch, alarm limit, and alarm priority is password protected. The monitor’s local password is required to change alarm settings.

[

User Password

]: changing alarm ON/OFF switch, alarm limit, and alarm priority is password protected. The user name and password saved in the MLDAP server are required to change alarm settings.

3. Select [

Arrh. Alarm Settings

] to set password for changing arrhythmia settings.

[

No Password

]: changing arrhythmia settings is not password protected.

[

Local Password

]: changing arrhythmia ON/OFF switch, alarm priority, and arrhythmia threshold is password protected. The monitor’s local password is required to change arrhythmia settings.

[

User Password

]: changing arrhythmia ON/OFF switch, alarm priority, and arrhythmia threshold is password protected. The user name and password saved in the MLDAP server are required to change arrhythmia settings.

4. Select [

User Maintenance

] to set password for changing user maintenance settings.

[

Local Password

]: changing user maintenance settings is password protected. The monitor’s local password is required to change user maintenance settings.

[

User Password

]: changing user maintenance settings is password protected. The user name and password saved in the MLDAP server are required to change user maintenance settings.

Passport 12m/Passport 17m Operator's Manual 31-13

To change the monitor’s local password for changing alarm settings and arrhythmia settings, select [

Modify Local

Alarm Password>>

], enter the new password, and select [

OK

].

To change the monitor’s local password for accessing the [

User Maintenance

] menu, select [

Modify Local User

Maintenance Password>>

], enter the new password, and select [

OK

].

31.8

Using DVI-VGA Adapter Box

The monitor can be connected with a VGA device via a DVI-VGA adapter box.

1. Connect the monitor’s DVI output with DVI-VGA adapter box’s DVI input.

2. Connect the DVI-VGA adapter box’s VGA output with VGA device.

31-14 Passport 12m/Passport 17m Operator's Manual

32

Batteries

32.1

Overview

This monitor is designed to operate from rechargeable Lithium-ion battery power when AC power is not available. It is recommended to always install fully charged batteries into the monitor to ensure normal monitoring in case of accidental power failure. If the AC power is interrupted during patient monitoring, the monitor automatically runs from the batteries.

On-screen battery symbols indicate the battery status as follows:

Indicates full battery capacity.

Indicates that the batteries work correctly. The solid portion represents the current charge level of the battery in proportion to its maximum charge level.

Indicates that the batteries have a low charge level and need to be charged.

Indicates that the batteries are almost depleted and need to be charged immediately.

Indicates that no batteries are installed or only one battery is installed to the Passport 17m monitor.

The capacity of the battery is limited. When the battery is low, the technical alarm area displays [

Low Battery

], the alarm lamp flashes, and monitor produces an alarm sound.

If the battery is depleted, the battery symbol on the screen flashes, the technical alarm area displays [

Battery

Depleted

], the alarm lamp flashes, and the monitor produces alarm sound. Connect the equipment to AC mains to run the equipment and charge the battery. Otherwise the monitor will shut down.

Passport 12m/Passport 17m Operator's Manual 32-1

32.2

Safety

CAUTION

Keep the battery out of children’s reach.

Use only specified batteries.

Keep the batteries in their original package until you are ready to use them.

Do not expose batteries to liquid.

High ambient temperature shortens battery run time.

Extremely high ambient temperature may cause battery overheat protection, and then the monitor will turn off.

If the battery shows signs of damage or signs of leakage, replace it immediately. Do not use a faulty monitor battery.

Do not crush, drop or puncture the battery. Mechanical abuse can lead to internal damage and internal short circuits. If a battery has been dropped or banged against a hard surface, whether damage is externally visible or not, remove the battery from use and dispose of it properly.

Please replace your battery when it reaches the end of its service life. Failure to replace the battery may cause serious damage to your device from battery overheating.

NOTE

Charge the batteries in the monitor.

Use AC power supply when iView is in use.

32.3

Battery Guidelines

Battery life expectancy depends on how frequent and how long it is used. For a properly maintained and stored lithium ion battery, its life expectancy is about 3 years. For more aggressive use models, life expectancy can be less. We recommend replacing lithium ion batteries after 500 full charge/discharge cycles or every 3 years from first use, whichever occurs first.

To get the most out of the battery, observe the following guidelines:

 The battery performance test must be performed every six months, before monitor repairs, or whenever poor battery performance is suspected.

Condition the batteries regularly to maintain their useful life. For details, refer to

32.6 Conditioning a Battery.

Remove the battery before the monitor is shipped or will not be used for more than two weeks. Leaving the battery in a monitor that is not in regular use will shorten the life of the battery.

32-2 Passport 12m/Passport 17m Operator's Manual

 The shelf life of a lithium Ion battery is about 6 months when the battery is stored with the battery power being

50% of the total power. In 6 months the battery power must be depleted before the Lithium Ion battery is fully charged. Then run the monitor on this fully charged battery .When its battery power becomes 50% of the total power, take out the battery from the monitor and store it.

Store batteries properly. For details, refer to

32.8 Storing a Battery

.

32.4

Installing or Replacing a Battery

Passport 12m

When the monitor uses two battery packs, one battery pack can easily be exchanged while the monitor operates from the other. If the monitor uses one battery pack, you should insert a new battery pack before the old one depletes.

To install or replace a battery:

1. Push down the button on the battery door and then slide backward as indicated to open the battery door.

2. Push aside the latch securing the battery and then remove the battery.

3. Place the new battery into the slot.

4. If necessary, replace the other battery following the steps above.

5. Restore the latch to the original position and close the battery door.

NOTE

Using two batteries are recommended when a SMR is connected.

Passport 17m

The monitor uses two battery packs. If the two batteries have different charge capacity, the message [

Diff. Battery

Voltages

] is displayed. In this case, apply AC power to the monitor until the two batteries have approximately equal charge capacity or are both fully charged. You cannot use them before they have approximately equal charge capacity or are fully charged. In situations where no patient monitoring is being performed or interrupting the patient monitoring is permitted, you can replace the batteries.

Passport 12m/Passport 17m Operator's Manual 32-3

When the monitor is operating on battery power, make sure the monitor is powered off before replacing a battery.

The monitor uses two batteries. You can install the batteries by following this procedure:

1. Turn off the patient monitor and disconnect the power cord and other cables.

2. Place the monitor with its face up.

3. Open the battery compartment door.

4. Place the batteries into the slots.

5. Close the battery door and place the monitor upright.

32.5

Charging a Battery

To optimize performance, a fully (or nearly fully) discharged battery should be charged as soon as possible. The battery is charged regardless of whether or not the monitor is currently turned on.

To charge the battery:

1. Install the battery in the monitor

2. Connect the monitor with the AC power source.

When the battery is charging, the AC power indicator and battery indicator are both on. If the monitor is powered on, the battery status symbol on the monitor screen displays when the charging is complete.

32.6

Conditioning a Battery

The performance of rechargeable batteries may deteriorate over time. If a battery is not conditioned for a prolonged time, its charge indication may not be accurate and you may wrongly evaluate the remaining battery runtime.

Keeping the battery continuously fully charged without conditioning will speed up battery aging and shorten its life time. You should condition a battery every three months, before it is used for the first time, or when its run time becomes noticeably shorter.

A battery conditioning cycle is one uninterrupted charge of the battery, followed by an uninterrupted battery discharge and charge.

32-4 Passport 12m/Passport 17m Operator's Manual

NOTE

The battery charge level indicator does not indicate the capacity or operating time of the battery. It only indicates the current battery charge level. The actual battery capacity decreases over time with the use of the battery. For an old battery, its capacity and operating time may not fulfill battery specifications even when the battery charge level indicates the battery isfully charged. Please replace the battery if its operating time is significantly lower than the specified time.

To condition a battery:

1. Disconnect the monitor from the patient and stop all monitoring and measuring procedures.

2. Insert the battery to be conditioned in the monitor battery slots.

3. Apply AC power to the monitor and allow the battery to charge uninterruptedly for 10 hours.

4. Remove AC power and allow the monitor to run from the battery until it shuts off.

5. Apply AC power again to the monitor and allow the battery to charge uninterruptedly for 6 hours.

6. This battery is now conditioned and the monitor can be returned to service.

CAUTION

Do not use the monitor to monitor the patient during battery conditioning.

NOTE

Do not interrupt battery conditioning.

32.7

Checking Battery Performance

The performance of a rechargeable battery will deteriorate over time. The battery performance test should be performed every six months, before monitor repairs, or whenever the battery performance is diminished. To check the battery performance:

1. Disconnect the monitor from the patient and stop all monitoring and measuring procedures.

2. Turn off the monitor.

3. Apply AC power to the monitor and allow the battery to charge uninterruptedly for 6 hours.

4. Remove AC power and allow the monitor to run from the battery until it shuts off.

5. Record the battery operating time.

The battery operating time directly reflects its performance. If the operating time of a battery is noticeably shorter than that stated in the specifications, replace the battery or contact the Mindray service personnel.

Passport 12m/Passport 17m Operator's Manual 32-5

NOTE

Battery operating time depends on the device configuration and operation. For example, high display brightness or measuring NIBP repeatedly will shorten the battery operating time.

32.8

Storing a Battery

When storing batteries, make sure that the battery terminals do not come into contact with metallic objects. If batteries are stored for an extended period of time, they should be placed in a cool place with a partial charge of 40% to 60% capacity.

Stored batteries should be conditioned every

6

months. Refer to section

32.6 Conditioning a Battery

for details.

NOTE

Remove the battery from the equipment if the equipment is not used for a prolonged period of time (for example, several weeks). Otherwise the battery may be over discharged and damaged.

Storing batteries at high temperature for an extended period of time will significantly shorten their life expectancy.

Storing batteries in a cool place will slow the aging process. Ideally the batteries should be stored at 15 °C.

Do not store the batteries in an environment above 60 °C or lower than -20 °C.

32.9

Recycling a Battery

Remove the old battery from the monitor and recycle it properly. Follow local laws for proper battery disposal.

WARNING

Do not open batteries, heat above 60 °C, incinerate batteries, or short the battery terminals. They may ignite, explode, leak, or heat up, causing personal injury.

32-6 Passport 12m/Passport 17m Operator's Manual

33

Care and Cleaning

33.1

General Points

In this chapter we only describe cleaning and disinfection of the main unit, modules, satellite module rack (SMR) and certain accessories. For the cleaning and disinfection of other reusable accessories, refer to instructions for use of the corresponding accessories.

Keep your equipment and accessories clean. To avoid damage to the equipment, follow these guidelines:

Always follow the manufacturer’s instructions for each cleaning / disinfecting agent.

Do not immerse any part of the equipment or accessories into liquid.

Do not pour liquid onto the equipment or accessories.

 Do not allow liquid to enter the case.

 Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners (such as acetone or acetone-based cleaners).

WARNING

Be sure to turn off the system and disconnect all power cables from the outlets before cleaning the equipment.

CAUTION

If you spill liquid on the equipment or accessories, contact Mindray or your service personnel.

Avoid wetting the pins and metal parts of the main unit, modules, SMR or accessories during cleaning and disinfection.

Use only Mindray approved cleaners and disinfectants and methods listed in this chapter to clean or disinfect your equipment. Warranty does not cover damage caused by unapproved substances or methods.

We make no claims regarding the efficacy of the listed chemicals or methods as a means for controlling infection. For the method to control infection, consult your hospital’s Infection Control Officer or

Epidemiologist.

The responsible hospital or institution shall carry out all cleaning and disinfection procedures specified in this chapter.

Refer to the respective instructions for use of the cleaning agents and disinfectants.

Do not mix disinfecting solutions (such as bleach and ammonia) as hazardous gases may result.

NOTE

To clean or disinfect reusable accessories, refer to the instructions delivered with the accessories.

Passport 12m/Passport 17m Operator's Manual 33-1

33.2

Cleaning and Disinfecting the Main Unit/Module/SMR

33.2.1

Approved Cleaning and Disinfecting Agents

The following table lists approved cleaning and disinfecting agents:

Product Name Product Type

Sodium hypochlorite bleach

*

Liquid

Hydrogen peroxide

*

Isopropanol

*

1-Propanol

*

Metrex CaviCide1™

Active Ingredients

Sodium hypochlorite bleach 0.5%

Hydrogen peroxide 3%

Isopropanol 70%

1-Propanol 50%

Virex

II 256

Virex

TB

Rely+On

TM

Virkon Ⓡ Powder

*

(Used as 1% solution)

*Perform® Classic

Concentrate OXY, 0.5%

Alpet

D2 Surface Sanitizing

Wipes

Powder

Wipes

Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0.28%, Isopropanol 17.2%

Didecyl dimethyl ammonium chloride 8.704%, n-Alkydimethyl benzyl ammonium chloride 8.190% n-Alkyl dimethyl benzyl ammonium chlorides 0.105%, n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.105%

Used as 1% solution

Biocidal active:

Pentapotassium bis (peroxymonosulphate) bis

(sulphate)(500g/kg),

Contains dipotassium peroxodisulphate.

Bactericidal 0.5% (5 g/l)

Fungicidal 1% (10 g/l)

Sporicidal 1% (10 g/l)

Enveloped virucidal (BVDV, Vaccinia) * 0.5% (5 g/l)

Isopropyl Alcohol 58.6000%

Octyl Decyl Dimethyl Ammonium chloride 0.0075%,

Dioctyl Dimethyl Ammonium Chloride 0.0030%

Sodium Hypochlorite 0.65% Clorox Dispatch Ⓡ Hospital

Cleaner Disinfectant Towels with Bleach

Clorox Healthcare Ⓡ Bleach

Germicidal Wipes

Clorox Healthcare

Hydrogen Peroxide Cleaner

Disinfectant Wipes

Diversey Oxivir

TB Wipes

Metrex CaviWipes™

Sodium Hypochlorite 0.55%

Hydrogen Peroxide 1.4%

PDI Sani-Cloth

AF3

Germicidal Disposable Wipe

PDI Sani-Cloth

Bleach

Germicidal Disposable Wipe

Wipes

Hydrogen Peroxide 0.5%

Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0.28%,

Isopropanol 17.2% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.14%, n-Alkyl dimethyl benzyl ammonium chlorides 0.14%

Sodium Hypochlorite 0.63%,other ingredients 99.37%

33-2 Passport 12m/Passport 17m Operator's Manual

Product Name

PDI Sani-Cloth

HB

Germicidal Disposable Wipe

PDI Sani-Cloth

Plus

Germicidal Disposable Cloth

PDI Super Sani-Cloth

Germicidal Disposable Wipe

Product Type

Wipes

VIRAGUARD Hospital Surface

Disinfectants

NOTE

Active Ingredients

n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.07%, n-Alkyl dimethyl benzyl ammonium chlorides 0.07% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.125%, n-Alky dimethyl benzyl ammonium chlorides 0.125% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.25%, n-Alkyl dimethyl benzyl ammonium chlorides 0.25%,

Isopropyl Alcohol 55.0%,

Isopropanol 70%,Other ingredients 30%

For equipment with the symbol , all the listed cleaning and disinfecting agents are available for use.

For equipment without the symbol available for use.

, only the cleaning and disinfecting agents marked with “*” are

33.2.2

Cleaning the Main Unit/Module/SMR

Your equipment shoul be cleaned on a regular basis. Before cleaning the equipment, consult your hospital’s regulations for cleaning the equipment.

To clean your equipment, follow this procedure:

1. Shut down the monitor and disconnect it from the AC power.

2. Clean the display screen with wipes or a soft cloth moistened with one of the cleaning agents listed in

33.2.1

Approved Cleaning and Disinfecting Agents

only.

3. Clean the exterior surface of the equipment, module or SMR with wipes or a soft cloth moistened with one of the

cleaning agents listed in

33.2.1 Approved Cleaning and Disinfecting Agents

only.

4. Wipe off all the cleaning agent residue with a dry cloth.

5. Dry your equipment in a ventilated, cool place.

33.2.3

Disinfecting the Main Unit/Module/SMR

Disinfect the equipment as required in your hospital’s servicing schedule using the disinfecting agents listed in the table above. Cleaning equipment before disinfecting is recommended.

CAUTION

Never use EtO or formaldehyde for disinfection.

Passport 12m/Passport 17m Operator's Manual 33-3

33.3

Cleaning and Disinfecting the Accessories

For the NIBP air hose, Masimo SpO

2

cable, Nellcor SpO

2

cable and NMT accessories, you should clean and disinfect them using the cleaners and disinfectants and methods listed in this section. For other accessories, you should consult the instructions delivered with the accessories.

CAUTION

Fluids entering the NIBP air hose can damage the equipment. When cleaning and/or disinfecting the NIBP air hose, prevent liquid from entering the hose.

Periodically inspect the NIBP air hose and connector for signs of wear or deterioration. Replace the NIBP air hose if you detect a leak. Dispose of damaged NIBP air hose according to local laws for disposal of hospital waster.

Never immerse or soak the accessories in any liquid.

Never clean or disinfect the metallic connectors at either end of the accessories.

Use only Mindray approved cleaners and disinfectants and methods listed in this section to clean or disinfect the accessories. Warranty does not cover damage caused by unapproved substances or methods.

To avoid long term damage, the accessories should be disinfected only when necessary as determined by your hospital’s policy.

33.3.1

Approved Accessories Cleaning and Disinfecting Agents

The following table lists approved NIBP air hose cleaning and disinfecting agents:

Product Name

Isopropanol

1-Propanol

Product Type

Liquid

Active Ingredients

Isopropanol 70%

1-Propanol 50%

Metrex CaviCide1™

Virex

TB

Rely+On TM Virkon

Powder

(Used as 1% solution)

Alpet Ⓡ D2 Surface Sanitizing

Wipes

Powder

Wipes

Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0.28%,

Isopropanol 17.2% n-Alkyl dimethyl benzyl ammonium chlorides 0.105%, n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.105%

Used as 1% solution

Biocidal active:

Pentapotassium bis (peroxymonosulphate) bis (sulphate)

(500g/kg),

Contains dipotassium peroxodisulphate 1%

Isopropyl Alcohol 58.6000%

Octyl Decyl Dimethyl Ammonium chloride 0.0075%,

Dioctyl Dimethyl Ammonium Chloride 0.0030%

33-4 Passport 12m/Passport 17m Operator's Manual

Product Name

Clorox Dispatch

Hospital

Cleaner Disinfectant Towels with Bleach

Metrex CaviWipes™

Product Type

Wipes

Active Ingredients

Sodium Hypochlorite 0.65%

Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0.28%,

Isopropanol 17.2%

PDI Sani-Cloth

AF3

Germicidal Disposable Wipe

PDI Sani-Cloth Ⓡ Plus

Germicidal Disposable Cloth n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.14%, n-Alkyl dimethyl benzyl ammonium chlorides 0.14% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.125%, n-Alky dimethyl benzyl ammonium chlorides 0.125% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.25%,

PDI Super Sani-Cloth

Germicidal Disposable Wipe n-Alkyl dimethyl benzyl ammonium chlorides 0.25%,

Isopropyl Alcohol 55.0%,

Isopropanol 70%,Other ingredients 30% VIRAGUARD Hospital Surface

Disinfectants

The following table lists approved Masimo SpO

2 cable cleaning and disinfecting agents:

Product Name Product Type Ingredients

Isopropanol Liquid Isopropanol 70%

The following table lists approved Nellcor SpO

2 cable cleaning and disinfecting agents:

Product Name Product Type

Sodium hypochlorite bleach Liquid

Isopropanol

Ingredients

Sodium hypochlorite bleach 0.5%

Isopropanol 70%

1-Propanol 1-Propanol 50%

Virex Ⓡ TB

Rely+On TM Virkon

Powder

(Used as 1% solution)

Powder

Wipes n-Alkyl dimethyl benzyl ammonium chlorides 0.105%, n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.105%

Used as 1% solution

Biocidal active:

Pentapotassium bis (peroxymonosulphate) bis (sulphate)

(500g/kg),

Contains dipotassium peroxodisulphate 1%

Sodium Hypochlorite 0.65%

Clorox Dispatch

Hospital

Cleaner Disinfectant Towels with Bleach

Clorox Healthcare Ⓡ Bleach

Germicidal Wipes

Clorox Healthcare

Hydrogen Peroxide Cleaner

Disinfectant Wipes

Diversey Oxivir

TB Wipes

PDI Super Sani-Cloth

Germicidal Disposable Wipe

Sodium Hypochlorite 0.55%

Hydrogen Peroxide 1.4%

Hydrogen Peroxide 0.5% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.25%, n-Alkyl dimethyl benzyl ammonium chlorides 0.25%,

Isopropyl Alcohol 55.0%,

Passport 12m/Passport 17m Operator's Manual 33-5

Product Name Product Type Ingredients

VIRAGUARD Hospital Surface

Disinfectants

Isopropanol 70%,

Other ingredients 30%

The following table lists approved NMT accessory cleaning and disinfecting agents:

Product Name Product Type

Sodium hypochlorite bleach Liquid

Isopropanol

Ingredients

Sodium hypochlorite bleach 0.5%

Isopropanol 70%

1-Propanol 1-Propanol 50%

Metrex CaviCide1™

Virex

Rely+On

TM

Virkon Ⓡ Powder

(Used as 1% solution)

Alpet

Wipes

TB

D2 Surface Sanitizing

Powder

Wipes

Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0.28%,

Isopropanol 17.2% n-Alkyl dimethyl benzyl ammonium chlorides 0.105%, n-Alkyl dimethyl ethylbenzyl ammonium chlorides

0.105%

Used as 1% solution

Biocidal active:

Pentapotassium bis (peroxymonosulphate) bis (sulphate)

(500g/kg),

Contains dipotassium peroxodisulphate 1%

Isopropyl Alcohol 58.6000%,

Octyl Decyl Dimethyl Ammonium chloride 0.0075%,

Dioctyl Dimethyl Ammonium Chloride 0.0030%

Sodium Hypochlorite 0.65% Clorox Dispatch

Hospital

Cleaner Disinfectant Towels with Bleach

Clorox Healthcare

Bleach

Germicidal Wipes

Clorox Healthcare

Hydrogen Peroxide Cleaner

Disinfectant Wipes

Diversey Oxivir Ⓡ TB Wipes

Metrex CaviWipes™

Sodium Hypochlorite 0.55%

Hydrogen Peroxide 1.4%

PDI Sani-Cloth

AF3

Germicidal Disposable Wipe

Hydrogen Peroxide 0.5%

Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0.28%,

Isopropanol 17.2% n-Alkyl dimethyl ethylbenzyl ammonium chlorides

0.14%, n-Alkyl dimethyl benzyl ammonium chlorides 0.14%

Sodium Hypochlorite 0.63%,other ingredients 99.37%

PDI Sani-Cloth

Bleach

Germicidal Disposable Wipe

PDI Sani-Cloth

Plus

Germicidal Disposable Cloth

PDI Super Sani-Cloth

Germicidal Disposable Wipe

Wipes n-Alkyl dimethyl ethylbenzyl ammonium chlorides

0.125%, n-Alky dimethyl benzyl ammonium chlorides 0.125% n-Alkyl dimethyl ethylbenzyl ammonium chlorides

0.25%,

33-6 Passport 12m/Passport 17m Operator's Manual

Product Name

VIRAGUARD Hospital Surface

Disinfectants

Product Type Ingredients

n-Alkyl dimethyl benzyl ammonium chlorides 0.25%,

Isopropyl Alcohol 55.0%,

Isopropanol 70%,

Other ingredients 30%

33.3.2

Cleaning the Accessories

You should clean the accessories (NIBP air hose, Masimo SpO

2

cable, Nellcor SpO

2

cable and NMT accessories) on a regular basis. Before cleaning the accessories, consult your hospital’s regulations for cleaning the accessories.

To clean the accessories (NIBP air hose, Masimo SpO

2

cable, Nellcor SpO

2

cable and NMT accessories), follow this procedure:

1. Clean the accessories with wipes or a soft cloth moistened with one of the cleaning agents listed in

33.3.1

Approved Accessories Cleaning and Disinfecting Agents

only.

2. Wipe off all the cleaning agent residue with a dry cloth.

3. Allow the accessories to air dry.

33.3.3

Disinfecting the Accessories

We recommend that the accessories (NIBP air hose, Masimo SpO

2

cable, Nellcor SpO

2

cable and NMT accessories) shoud be disinfected only when necessary as determined by your hospital’s policy, to avoid long term damage to the accessories. Cleaning the accessories before disinfecting is recommended.

33.4

Sterilization

Sterilization is not recommended for this equipment, related products, accessories, or supplies unless otherwise indicated in the Instructions for Use that accompany the products, accessories or supplies.

Passport 12m/Passport 17m Operator's Manual 33-7

FOR YOUR NOTES

33-8 Passport 12m/Passport 17m Operator's Manual

34

User Maintenance

CAUTION

Failure on the part of the responsible individual hospital or institution using this equipment to implement a recommended maintenance schedule may cause undue equipment failure and possible health hazards.

The safety checks or maintenance involving any disassembly of the equipment should be performed by professional servicing personnel. Otherwise, undue equipment failure and possible health hazards could result.

No modification of this equipment is allowed.

Do not open the equipment housings. All servicing and future upgrades must be carried out by the service personnel trained and authorized by Mindray only.

If you discover a problem with any of the equipment, contact your service personnel or Mindray.

The service personnel must be properly qualified and thoroughly familiar with the operation of the equipment.

34.1

General Inspection

A thorough inspection should be performed by qualified service personnel to ensure the reliability before first use, after your monitor has been used for 12 months, or whenever your monitor is repaired or upgraded.

Follow these guidelines when inspecting the equipment:

 Make sure that the environment and power supply meet the specifications.

 Inspect the equipment and its accessories for mechanical damage.

 Inspect all power cords for damage, and make sure that their insulation is in good condition.

 Make sure that only specified accessories are applied.

 Inspect if the alarm system functions correctly.

 Make sure that the recorder functions correctly and the recorder paper meets the specifications.

Make sure that the batteries meet the performance specifications.

Make sure that the monitor is in good working condition.

In case of any damage or abnormity, do not use the monitor. Contact the hospital’s biomedical engineers or your service personnel immediately.

Passport 12m/Passport 17m Operator's Manual 34-1

34.2

Maintenance and Testing Schedule

The following maintenance and tests, except for visual inspection, power on test, touchscreen calibration, battery check,

ECG calibration, NIBP leakage, CO

2

leakage, AG leakage, and recorder check, should be carried out by the service personnel only. Ensure the monitor is safe and the performance tested by qualified service personnel before initial installation, after upgrade, or during regular scheduled maintenance. Contact your service personnel if any maintenance is required. Make sure to clean and disinfect the equipment before any test and maintenance.

CAUTION

Changing the settings in [User Settings >>] and [Factory Maintenance >>] menus may cause data loss.

Service personnel should acquaint themselves with the test tools and make sure that the test tools and cables are applicable.

Check/Maintenance Item

Performance Tests

Visual inspection

ECG test

Performance test

Verification

Recommended Frequency

When first installed or reinstalled.

Resp performance test

SpO

2

test

Pressure check

NIBP test

Leakage test

Temp test

IBP performance test

C.O. test

Sidestream and

Microstream CO

2

tests and calibration

Leakage test

Performance test

AG test

Calibration

Leakage test

Performance test

Calibration

BIS test

RM test

Interconnecting function

CCO/SvO

2

test

NMT test

Output calibration

Performance test

Sensor Check

ScvO

2 test

Nurse call relay performance test

Analog output performance test

34-2

1. If the user suspects that the measurement is incorrect.

2. Following any repairs or replacement of relevant module.

3. At least once every two years.

Note: At least once a year is recommended for NIBP, CO

2,

NMT and

AG.

If the user suspects that the analog output does not work well.

Passport 12m/Passport 17m Operator's Manual

Check/Maintenance Item

Electrical Safety Tests

Electrical safety test

Other Tests

Recommended Frequency

At least once every two years

Power on test

Touchscreen calibration

Recorder check

Network print test

Device integration check

Battery check

Functionality test

Performance test

1. When first installed or reinstalled.

2. Following any maintenance or the replacement of any main unit parts.

1. When the touchscreen appears abnormal.

2. After the touchscreen is replaced.

Following any repair or replacement of the recorder.

1. When first installed.

2. Whenever the printer is serviced or replaced.

1. When first installed.

2. Following any repair or replacement of the external device.

1. When first installed.

2. Whenever a battery is replaced.

Once a year or if the battery run time reduced significantly.

34.3

Checking Monitor and Module Information

To view the information about system start time, selftest, etc., select [

Main Menu

]

[

Maintenance >>

]

[

Monitor

Information >>

]. You can print out the information for the convenience of troubleshooting. The information will not be saved during shut down.

You can also view the information about the monitor configuration and system software version by selecting [

Main

Menu

]

[

Maintenance >>

]

[

Software Version >>

].

34.4

ECG Verification

The ECG signal may be inaccurate due to hardware or software problems. As a result, the ECG wave amplitude becomes greater or smaller. In that case, you need to verify the ECG module.

1. Select the ECG parameter window or waveform area

[

Filter

]

[

Diagnostic

].

2. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password

[

Module

Maintenance >>

]

[

Calibrate ECG

]. A square wave appears on the screen and the message [

ECG Calibrating

] is displayed.

3. Compare the amplitude of the square wave with the wave scale. The difference should be within 5%.

4. After the verification is completed, select [

Stop Calibrating ECG

].

You can print the square wave and wave scale and then measure the difference between them if necessary. If the difference exceeds 5%, contact your service personnel.

Passport 12m/Passport 17m Operator's Manual 34-3

34.5

NIBP Leakage Test

The NIBP leakage test checks the integrity of the system and of the valve. It is required at least once a year or when you doubt the measured NIBP. If the test failed, corresponding prompt messages will be given. If no message is displayed, it means no leakage is detected.

Tools required:

An adult cuff

Air tubing

 A correct sized cylinder

To perform a leakage test:

1.

Set the patient category to [

Adu

].

2.

Connect the cuff to the NIBP connector on the monitor.

3.

Wrap the cuff around the cylinder as shown below.

Monitor

Connector for

NIBP cuff

Air tubing

Cylinder

Cuff

4.

Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]. Enter the required password and then select

[

Ok

].

5.

Select [

Module Maintenance >>

]

[

NIBP Leakage Test

]. The NIBP display shows [

Leakage Testing…

].

After about 20 seconds, the monitor will automatically deflate. This means the test is completed. If the message [

NIBP

Pneumatic Leak

] is displayed, it indicates that the NIBP airway may have leakages. Check the tubing and connections for leakages. If you ensure that the tubing and connections are all correct, perform a leakage test again.

If the problem persists, contact your service personnel.

NOTE

The leakage test is intended for use to simply determine whether there are leakages in the NIBP airway. It is not the same as that specified in the EN 1060-3 standard.

34-4 Passport 12m/Passport 17m Operator's Manual

34.6

CO

2

Leakage Test

For sidestream and microstream CO

2 modules, leakage test is needed every year or when you suspect the measurement.

To perform this test:

1. Connect the CO

2

module with the patient module.

2. Wait until CO

2

warmup is finished and then use your hand or other objects to completely block the gas inlet of the module or watertrap. The sidestream and microstream CO

2

modules will behave as follows:

Sidestream: The alarm message [

CO

2

FilterLine Err

] is displayed on the screen after certain time. Block the gas inlet for another 30 s. If the alarm message does not disappear, it indicates that the module does not leak.

Microstream: The alarm message [

CO

2

Purging

] is displayed on the screen after certain time. Block the gas inlet for another 30s. If alarm message [

CO

2

FilterLine Err

] is shown, it indicates that the module does not leak.

34.7

AG Leakage Test

The AG leakage test is required before every AG measurement. Follow this procedure to perform the test:

1. Plug the AG module into the module rack.

2. Wait until the AG module warmup is finished, and then use your hand or other objects to completely block the gas inlet of the AG module. An alarm message [

AG Airway Occluded

] will appear on the screen.

3. Block the gas inlet for 60 s. Select [

Main menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

→ enter the required password

[

Module Maintenance >>

]

[

Calibrate AG >>

], and check that current flow rate is less than 10ml/min.

If the flow rate is less than 10ml/min and the alarm message [

AG Airway Occluded

] remain displayed, it indicates that the module does not leak.

If the alarm message does not appear, or the flow rate is greater than or equal to 10ml/min, it indicates that the module leaks. If the problem remains, contact your service personnel for help.

34.8

Checking NMT Sensor

NMT sensor check is required once a year or when you doubt the measured values.

To check the NMT transducer,

1. Select [

Main Menu

]

[

Maintenance >>

]

[

User Maintenance >>

]

enter the required password

[Module

Maintenance>>]

[

NMT Sensor Check >>

].

2. Follow the on-screen instructions to check the NMT sensor in four ways.

If sensor check completes successfully, the message “Test passed. The function of NMT sensor is OK” is presented. If any of the four steps fails, check if the sensor is placed correctly as instructed, and perform the sensor check again. Replace

Passport 12m/Passport 17m Operator's Manual 34-5

the sensor or contact your service personnel if you cannot pass the sensor check.

NOTE

Stop NMT measurement or calibration before starting the NMT sensor check.

Take care when handling the the NMT sensor, avoiding rough impact.

34.9

Calibrating the Touchscreen

To calibrate the touchscreen:

1. Select [

Main Menu

]

[

Maintenance >>

]

[

Cal. Touchscreen

].The of the screen.

symbol will appear at different positions

2. Select each as it appears on the screen.

3. After the calibration is completed, the message [

Screen Calibration Completed!

] is displayed. Select [

Ok

] to confirm the completion of the calibration.

34-6 Passport 12m/Passport 17m Operator's Manual

35

Accessories

The accessories listed in this chapter comply with the requirements of IEC 60601-1-2 when in use with the patient monitor. The accessory material that contacts the patients has undertaken the bio-compatibility test and is verified to be in compliance with ISO 10993-1. For details about the accessories, refer to the instructions for use provided with the accessory.

WARNING

Use accessories specified in this chapter. Using other accessories may cause damage to the patient monitor or not meet the claimed specifications.

Single-use accessories are not designed to be reused. Reuse may cause a risk of contamination and affect the measurement accuracy.

CAUTION

Check the accessories and their packages for any sign of damage. Do not use them if any damage is detected.

Use the accessories before the expiry date if their expiry date is indicated.

The disposable accessories shall be disposed of according to hospital's regulations.

NOTE

This manual describes all the accessories that are validated for use. Not all accessories are available in every market.

35.1

ECG Accessories

ECG Electrodes

Model

31499224

2245-50

/

/

Part No.

0010-10-12304

9000-10-07469

0681-00-0098-01

0681-00-0098-02

Description

ECG electrode, Covidien, 10 pieces

ECG electrode, 3M, (package of 50)

Radio Opaque, Pre-wired, 3-lead ECG electrodes,

AHA, 18" (45.7 cm), 3 pcs/pack, 100 packs/box

Radio Translucent, Pre-wired, 3-lead ECG electrodes, AHA, 18" (45.7 cm), 3 pcs/pack, 100 packs/box

Patient category

Adult

Pediatric

Neonate

Neonate

Passport 12m/Passport 17m Operator's Manual 35-1

ECG Cables

Model

EV6222

EV6201

EV6202

EV6203

EV6211

EV6212

EV6206

EV620

7

EV6216

EV6217

/

/

EV6208

ECG Leadwires

PN

040-000754-00

0010-30-42719

0010-30-42720

0010-30-42721

0010-30-42723

0010-30-42724

009-005266-00

009-005267-00

009-005268-00

009-005269-00

040-001416-00

009-003652-00

040-003528-00

Description

ECG cable, 12-pin, 3-lead, defibrillation-proof, DIN connector

ECG cable,12-pin, 3/5-lead, defibrillation-proof,

Applicable patient

Neonate, infant

Adult, pediatric

AHA/IEC

ECG cable,12-pin, 3-lead, defibrillation-proof, AHA/IEC Neonate, infant

ECG cable, 12-lead, defibrillation-proof, AHA

ECG cable, 12-pin, 3/5-lead, ESU-proof, AHA/IEC

ECG cable, 12-pin, 3-lead, ESU-proof, AHA/IEC

ECG cable, defibrillation-proof, 3.1 m, T/N series

Adult

Adult, pediatric

Neonate, infant

Adult, pediatric

ECG cable, defibrillation-proof, 6.2 m, T/N series

ECG cable, defibrillation-proof, 3.1 m, T/N series

ECG cable, defibrillation-proof, 6.2 m, T/N series

ECG cable, 12-pin 3/5-lead defibrillation-proof

ECG cable, 12-pin 3/5-lead, ESU-proof

12Pin 12-Lead Host Cable, AHA, Def-P(MW)

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

3-Electrode Leadwires

Model

EL6305A

EL6301A

EL6303A

EL6301B

EL6311B

/

/

/

/

/

/

/

/

/

EY6316B

EY6305B

EY6316A

EY6305A

EY6310B

PN

0010-30-42896

0010-30-42726

Description

ECG leadwires, 3-lead, AHA, pinch

ECG leadwires, 3-lead, AHA, pinch

0010-30-42731

0010-30-42734

ECG leadwires, 3-lead, AHA, pinch, long

ECG leadwires, 3-lead, AHA, snap

040-000146-00 ECG leadwires, 3-lead, AHA, snap, single patient use

0012-00-1503-05 ECG Leadwires, 3-lead, Snap, 24" (61.0 cm), AAMI

0012-00-1503-06 ECG Leadwires, 3-lead, Snap, 36" (101.6 cm), AAMI

0012-00-1503-14 ECG Leadwires, 3-lead, Snap, 24" (61.0 cm), IEC

0012-00-1503-15 ECG Leadwires, 3-lead, Snap, 36" (101.6 cm), IEC

0012-00-1514-05 ECG Leadwires, 3-lead, Pinch, 24" (61.0 cm), AAMI

0012-00-1514-06 ECG Leadwires, 3-lead, Pinch, 36" (101.6 cm), AAMI

0012-00-1514-14 ECG Leadwires, 3-lead, Pinch, 24" (61.0 cm), IEC

0012-00-1514-15 ECG Leadwires, 3-lead, Pinch, 36" (101.6 cm), IEC

040-000748-01 Disp leadwires, 3-L snap, AAMI, 20 pk

009-004765-00 3-lead, new telemetry, AHA, snap, 24"

009-004766-00

009-004771-00

009-004772-00

009-004777-00

3-lead, new telemetry, AHA, snap, 36"

3-lead, new telemetry, AHA, pinch, 24"

3-lead, new telemetry, AHA, pinch, 36"

3-lead, new telemetry, disposable, AHA, snap, 24"

Applicable patient

Neonate, infant

Adult, pediatric

Adult, pediatric

Adult, pediatric

Neonate, infant

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

35-2 Passport 12m/Passport 17m Operator's Manual

Model

/

572A

4080

4081

583A

4092

4089

/

5-Electrode Leadwires

/

/

/

/

Model

EL6501A

EL6503A

EL6501B

/

/

/

/

/

EY6511B

EY6512B

EY6511A

EY6512A

EY6507B

12-Electrode Leadwires

Model PN

EL6801A

EL6803A

EL6801B

EL6803B

0010-30-42902

0010-30-42904

0010-30-42906

0010-30-42908

PN

0010-30-42727

0010-30-42729

0010-30-42735

Description

ECG leadwires, 5-lead, AHA, pinch

ECG leadwires, 5-lead, AHA, pinch, long

ECG leadwires, 5-lead, AHA, snap

0012-00-1503-02 ECG Leadwires, 5-lead, Snap, 24" (61.0 cm), AAMI

0012-00-1503-03 ECG Leadwires, 5-lead, Snap, 36" (101.6 cm), AAMI

0012-00-1503-11 ECG Leadwires, 5-lead, Snap, 24" (61.0 cm), IEC

0012-00-1503-12 ECG Leadwires, 5-lead, Snap, 36" (101.6 cm), IEC

0012-00-1514-02 ECG Leadwires, 5-lead, Pinch, 24" (61.0 cm), AAMI

0012-00-1514-03 ECG Leadwires, 5-lead, Pinch, 36" (101.6 cm), AAMI

0012-00-1514-11 ECG Leadwires, 5-lead, Pinch, 24" (61.0 cm), IEC

0012-00-1514-12 ECG Leadwires, 5-lead, Pinch, 36" (101.6 cm), IEC

040-000746-01

009-004782-00

009-004783-00

Disp leadwires, 5-L snap, AAMI, 20 pk

5-lead, new telemetry, AHA, snap, 24"

5-lead, new telemetry, AHA, snap, 36"

009-004786-00

009-004787-00

009-004790-00

5-lead, new telemetry, AHA, pinch, 24"

5-lead, new telemetry, AHA, pinch, 36"

5-lead, new telemetry, disposable, AHA, snap, 24"

Description

ECG leadwires, 12-lead,limb lead, AHA, pinch

ECG leadwires, 12-lead,chest lead, AHA, pinch

ECG leadwires, 12-lead,limb lead, AHA, snap

ECG leadwires, 12-lead,chest lead, AHA, snap

35.2

SpO

2

Accessories

Extension Cable

Part No.

115-020768-00

0010-20-42712

040-003378-00

040-003379-00

040-003310-00

040-003426-00

040-003381-00

040-003310-00

Description

8-pin, purple connector, 2.1 m, Masimo

8-pin, 2.5 m, Nellcor

RD SET MD 14-05, PC 5 ft

RD SET MD 14-12, patient cable 12 ft

8pin Masimo cable (RD SET) for MAHWAH

LNCS to RD adapter

RD to LNC Adapter Cable

8-pin Masimo Cable

Applicable patient

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Applicable patient

Adult

Adult

Adult

Adult

Applicable patient

All

All

All

All

All

All

All

All

Passport 12m/Passport 17m Operator's Manual 35-3

SpO

2

Sensors

Masimo SpO

2

Module

Model

LNCS DCI

LNCS DCIP

LNCS NeoPt

LNCS Neo

LNCS Inf

PN

0600-00-0126

0600-00-0158

Description

Reusable SpO

2

sensor

0600-00-0127 Reusable SpO

2

sensor

0600-00-0156 Disposable SpO

2

sensor

0600-00-0157 Disposable SpO

2

sensor

Disposable SpO

2

sensor

Applicable patient

Adult

Pediatric

Neonate

Adult (> 40 kg)

Neonate (< 3 kg)

Infant neonate ( 3 to 20 kg)

Application site

Finger

Finger

Foot

Finger

Foot

Toe

Foot

LNCS Pdtx

LNCS Adtx

4050

4051

4053

0600-00-0122

0600-00-0121

Disposable SpO

2

sensor

Disposable SpO

2

sensor

Pediatric

Adult

040-003376-00 RD SET DCI, adult reusable sensor Adult (> 30 kg)

RD SET DCI, pediatric reusable

040-003377-00

Pediatric (10 to 50 kg) sensor

040-003380-00

RD Set TC-I SpO

2

reusable tip-clip ear sensor, 3ft

Adult (> 30 kg)

Finger

Finger

Finger

Finger

Ear

4000

4001

040-003382-00 RD SET adhesive disposable sensor Adult (> 30 kg)

040-003383-00

RD SET PDT adhesive disposable senso

Pediatric (10 to 50 kg)

Finger

Finger

4002

4003

4004

4005

040-003384-00

040-003385-00

040-003386-00

040-003387-00

RD Set infant adhesive disposable sensor

RD Set neo adhesive disposable sensor

RD Set NeoPt adhesive disposable sensor

RD Set NeoPt-500 nonadhesive disposable sensor

Infant (3 to 20 kg)

Adult (> 40 kg)

Neonate (< 3 kg)

Neonate (< 1 kg)

Neonate (< 1 kg)

Toe/Finger

Foot/Finger

Foot

Foot

Wavelength of Masimo SpO

2

sensors: red light: 660 nm; infrared light: 940 nm.

The maximum photic output consumption of the sensor is less than 18 mW.

The information about the wavelength range and maximum photic output consumption can be especially useful to clinicians (for example, when photodynamic therapy is performed).

35.3

NIBP Accessories

NIBP hoses

Model

CM1901

CM1903

Part No.

6200-30-11560

6200-30-09688

Description

Reusable NIBP hose

Reusable NIBP hose

Applicable patient

Neonate

Adult, pediatric, infant

35-4 Passport 12m/Passport 17m Operator's Manual

Cuffs

Model

CM1500A

CM1500B

CM1500C

CM1500D

Part No.

001B-30-70692

001B-30-70693

001B-30-70694

001B-30-70695

/

/

/

/

/

/

/

/

/

/

/

/

/

/

/

115-027563-00

115-027564-00

115-027565-00

115-027566-00

115-027567-00

115-027568-00

115-027569-00

115-027713-00

115-027714-00

115-027715-00

115-027716-00

115-027717-00

115-027718-00

115-027719-00

115-031807-00

35.4

Temp Accessories

Temp Cables

Description

NIBP cuff, single patient use, size 1, 3.1 to 5.7 cm arm circumference, 20 pcs/box

NIBP cuff, single patient use, size 2, 4.3 to 8.0 cm arm circumference, 20 pcs/box

NIBP cuff, single patient use, size 3, 5.8 to 10.9 cm arm circumference, 20 pcs/box

NIBP cuff, single patient use, size 4, 7.1 to 13.1 cm arm circumference, 20 pcs/box

Disposable bladderless cuff, 10 - 19 cm, 10 pcs/box

Disposable bladderless cuff, 18 – 26 cm, 10 pcs/box

Disposable bladderless cuff, 24 – 35 cm, 10 pcs/box

Disposable bladderless cuff, 33 – 47 cm, 10 pcs/box

Disposable bladderless cuff, 46 – 66 cm, thigh, 5 pcs/box

Disposable bladderless long cuff, 24 – 35 cm, 10 pcs/box

Disposable bladderless long cuff, 33 – 47 cm, 10 pcs/box

Reusable bladderless cuff, 10 - 19 cm

Reusable bladderless cuff, 18 – 26 cm

Reusable bladderless cuff, 24 – 35 cm

Reusable bladderless cuff, 33 – 47 cm

Reusable bladderless cuff, 46 – 66 cm

Reusable bladderless long cuff, 24 – 35 cm

Reusable bladderless long cuff, 33 – 47 cm

Starter kit, reusable bladderless cuff

Model

MR420B

MR421

MR420B

Part No.

0011-30-37391

0010-30-43056

040-001235-00

Description

2-pin, extension cable, applicable

Adapter cable

Temperature adapter cable

2-pin plug

Applicable patient

All

All

All

Applicable patient

Neonate

Neonate

Neonate

Neonate

Child

Small adult

Adult

Large adult

Adult

Adult

Large adult

Child

Small adult

Adult

Large adult

Thigh

Adult

Large adult

/

Passport 12m/Passport 17m Operator's Manual 35-5

Temp Probes

/

/

/

/

/

Model

MR411

MR412

MR401B

MR402B

MR403B

MR404B

Part No.

0011-30-90446

0011-30-90447

0011-30-37392

0011-30-37394

0011-30-37393

0011-30-37395

Description

Disposable temperature probe, esophageal

Disposable temperature probe, skin

Reusable temperature probe, esophageal

Reusable temperature probe, esophageal

Reusable temperature probe, skin

Reusable temperature probe, skin

0206-03-0112-02 PROBE,D TEMP,ES400-12 (box of 20)

0206-03-0118-02 PROBE,D TEMP,ES400-18 (box of 20)

0206-03-0209-02 PROBE,D TEMP,ER 400-9 (box of 20)

0206-03-0212-02 PROBE,D TEMP,ER400-12 (box of 20)

0206-03-0300-02 PROBE,D TEMP,STS-400 (box of 20)

35.5

IBP/ICP Accessories

IBP Accessories

/

/

Model

IM2202

IM2201

Part No.

001C-30-70757

001C-30-70759

0010-20-42795

0010-21-12179

/

ICP Accessories

0010-21-43082

Description

12-pin IBP cable, BD

12-pin IBP cable, Abbott

IBP cable adapter (12 pin to 6 pin)

12 Pin IBP Cable for Edwards

12 Pin IBP Cable for Memscap (SP844 82031 transducer)

Model

/

ICT/B

Part No.

0010-30-42742

0010-10-12151

35.6

C.O. Accessories

Model

CO7702

SP4042

SP5045

MX387

/

/

93522

Part No.

0010-30-42743

6000-10-02079

6000-10-02080

6000-10-02081

0012-00-1519

6800-30-50617

0012-00-1520

Description

12-pin ICP cable, Gaeltec

Disposable ICP transducer, Gaeltec

Adult

Adult

Adult

Adult

Adult

Applicable patient

All

All

Adult

Pediatric, neonate

Adult

Pediatric, neonate

/

/

/

Applicable patient

All

All

Applicable patient

All

All

Description

12-pin C.O. cable.

TI sensor

TI sensor housing

12 cc control syringe W/1CC stop W/rotator

Cable, C.O. Edwards Lifesciences bath probe adapter

Becton Dickinson C.O. accessory kit (includes 12 pin cable, BD accessories)

In-line inject temperature Probe

35-6 Passport 12m/Passport 17m Operator's Manual

35.7

CO

2

Accessories

Sidestream CO

2

module

Model

4000

4100

4200

Part No.

M02A-10-25937

M02A-10-25938

M02B-10-64509

60-15200-00

60-15300-00

60-14100-00

9200-10-10533

9200-10-10555

9000-10-07486

040-001187-00 040-001187-00

60-13100-00 125-000005-00

60-13200-00 125-000006-00

100-000080-00 115-058733-00

100-000081-00 115-058734-00

Description

Nasal CO

Nasal CO

2

Nasal CO

2

2

sample cannula sample cannula

sample cannula

Applicable patient Applicable for

Adult

Pediatric

Neonate

Two-slot CO

2 module and one-slot CO

2 module

Sampling line, 2.5m, disposable Adult, pediatric

Sampling line, 2.5m, disposable Neonate

Airway adapter, disposable, straight Adult, pediatric

Airway adapter, disposable, elbow

Watertrap, 10pcs/box

Watertrap, 10pcs/box

DRYLINE II Watertrap, 10pcs/box

DRYLINE II Watertrap, 10pcs/box

Neonate

Adult, pediatric

Neonate

Adult, pediatric

Neonate

Two-slot CO module

One-slot CO

2 module

2

Passport 12m/Passport 17m Operator's Manual 35-7

35.8

AG Accessories

Model

60-13100-00

Part No.

125-000005-00

60-13200-00

60-14100-00

125-000006-00

9000-10-07486

040-001187-00 040-001187-00

60-15200-00 9200-10-10533

60-15300-00 9200-10-10555

/ 115-052162-00

100-000080-00 115-058733-00

100-000081-00 115-058734-00

60-14200-00 9000-10-07487

Description

Watertrap, 10pcs/box

Watertrap, 10pcs/box

Airway adapter, disposable, straight

Airway adapter, disposable, elbow

Sampling line, 2.5m, disposable

Sampling line, 2.5m, disposable

AG exhasust gas outlet assembly

DRYLINE II Watertrap, 10pcs/box

DRYLINE II Watertrap, 10pcs/box

Airway adapter, disposable,elbow

Applicable patient Applicable for

Adult, pediatric

Neonate

/

Three-slot AG module

Neonate

Adult, pediatric

Three-slot AG module and two-slot AG module

Neonate

Adult, pediatric

Adult, pediatric

Neonate

Adult, pediatric

Two-slot AG module

35.9

BIS Accessories

Model

186-0195-MR

186-0224-MR

Part No.

6800-30-50761

115-005707-00

Description

BIS module and cable

BISx4 module and cable

Applicable patient

/

/

35.10

NMT Accessories (for Mindray NMT module)

Model

NM13101

NM13401

NM13701

/

Part No.

040-001462-00

040-001463-00

040-001464-00

040-002258-00

Description

NMT cable

NMT sensor cable

NMT stimulation cable

NMT sensor securing strap,20 pcs/box

35.11

RM Accessories

Model

CP12901

PN155362

PN155500

Part No.

6800-20-50328

0010-30-42678

0010-30-42680

Description

RM connector

Flow sensor, Hamilton, reusable

Flow sensor, Hamilton, disposable

35.12

CCO/SvO

2

Accessories

Model

/

Part No.

009-000259-00

Description

CCO/SvO

2

Cable

Applicable patient

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Applicable patient

/

Adult, pediatric

Infant

Applicable patient

/

35-8 Passport 12m/Passport 17m Operator's Manual

35.13

ScvO

2

Accessories

Model

PC3030

PV2022-37

PV2022-35

Part No.

115-008191-00

040-000919-00

040-000920-00

Description

8Pin ScvO

2

Module and Cable

CeVOX Probe (37cm)

CeVOX Probe (35cm)

35.14

Installation Accessories

Passort 17m

Part No.

045-003426-00

045-000933-00

045-000934-00

045-000936-00

045-001302-00

Material

Main unit wall mount bracket

Satellite module rack wall mount bracket

Keyboard wall mount bracket

Display wall mount bracket

Passport 17m wall mount bracket (VHM)

Passport 12m

Part No.

045-000915-00

045-003426-00

045-000933-00

045-000934-00

045-000936-00

045-001275-00

045-001300-00

045-001301-00

Material

Roll stand

Main unit wall mount bracket

Satellite module rack wall mount bracket

Keyboard wall mount bracket

Display wall mount bracket

Passport 12m quick release

Passport 12m wall mount bracket

Passport 12m wall mount bracket (VHM)

/

/

Applicable patient

/

Passport 12m/Passport 17m Operator's Manual 35-9

35.15

Micellaneous Accessories

Part No.

022-000008-00

DA8K-10-14452

1000-21-00122

009-000027-00

8000-21-10361

115-004861-00

009-003648-00

009-003903-00

115-011465-00

023-000524-00

023-000525-00

023-000217-00

023-001673-00

023-000846-00

023-001566-00

023-001128-00

A30-000001---

115-046944-00

115-046949-00

115-033914-00

115-033911-00

Material

Lithium battery, LI23S002A

Power cord (America)

Grounding cable

Defibrillator synchronization cable

Nurse call cable (≤60W, ≤2A, ≤36VDC, ≤25VAC)

DVI-VGA adapter box

Cable protecting tube

Accessories management tape

BeneLikn ID adapter

Wireless mouse and keyboard kit

Wired mouse and keyboard kit

USB drive, 4GB, USB 2.0 (TRANSCEND)

USB drive, 8GB, USB 3.0 (APACER)

CF card

HP LaserJet Enterprise M608n

Elo ET1929LM, 19-inch black LCD

Recording paper, 50 mm*20 m

SMR material package(2M/handle/hook)

SMR material package(2M/no handle)

SMR NIBP bracket installation kit

SMR cable management hook installation kit

35-10 Passport 12m/Passport 17m Operator's Manual

36

Part Numbers for Canadian Use

All the part numbers listed in the table below are modules that are compatible with the Passport 12m and Passport

17m patient monitors. Our Passport 12m and Passport 17m patient monitors offer various optional modules to accommodate the customer's different needs.

Part No.

M51AF-PA00046

115-023313-00

M51AF-PA00050

M51AF-PA00049

Description

Masimo SpO

2

, 3/5-lead ECG, IBP

Contents

Module Part Number: 115-022715-00

Accessories kit (PN 115-023313-00) which contains:

040-001235-00 - Temperature Cable

040-001416-00 - ECG cable

0012-00-1503-02 - ECG Leadwire

115-020768-00 - Masimo SpO

2

cable

100-000078-00 - Reusable Adult Masimo SpO

2 sensor

115-027715-00 - Reusable Adult NIBP Cuff

6200-30-09688 - NIBP Hose

MPM Accessory kit - Masimo SpO

2

040-001235-00 - Temperature Cable

040-001416-00 - ECG cable

0012-00-1503-02 - ECG Leadwire

115-020768-00 - Masimo SpO

2

cable

100-000078-00 - Reusable Adult Masimo SpO

2 sensor

115-027715-00 - Reusable Adult NIBP Cuff

6200-30-09688 - NIBP Hose

Masimo SpO

2

, 3/5-lead ECG, IBP, Advanced

Arrhythmia and ST analysis

This kit has the same contents as M51AF-PA00046 with the difference between them being that this configuration contains the Advanced ST/ARR

Software

Module Part Number:115-022718-00

Accessories kit (PN 115-023313-00) which contains:

040-001235-00 - Temperature Cable

040-001416-00 - ECG cable

Masimo SpO

2

, 3/5/12-lead ECG, IBP, 0012-00-1503-02 - ECG Leadwire

Advanced Arrhythmia and ST analysis 115-020768-00 - Masimo SpO

2

cable

100-000078-00 - Reusable Adult Masimo SpO

2 sensor

115-027715-00 - Reusable Adult NIBP Cuff

6200-30-09688 - NIBP Hose

Passport 12m/Passport 17m Operator's Manual 36-1

Part No.

M51AF-PA00042

115-023314-00

M51AF-PA00044

M51AF-PA00045

115-030472-00

6800-30-50878

6800-30-50144

115-013249-00

115-030585-00

6800-30-50820

36-2

Description

Nellcor SpO

2

, 3/5-lead ECG, IBP

Contents

Module Part Number 115-022716-00

Accessories kit (PN 115-023314-00) which contains:

040-001235-00 - Temperature Cable

040-001416-00 - ECG cable

0012-00-1503-02 - ECG Leadwire

0010-20-42712 - Nellcor SpO

2

cable License

9000-10-05161 - Reusable Adult Nellcor SpO

2 sensor

115-027715-00 - Reusable Adult NIBP Cuff

6200-30-09688 - NIBP Hose

MPM Accessory kit - Nellcor SpO

Nellcor SpO

2

, 3/5/12-lead ECG, IBP, Advanced

Arrhythmia and ST analysis

2

040-001235-00 - Temperature Cable

040-001416-00 - ECG cable

0012-00-1503-02 - ECG Leadwire

0010-20-42712 - Nellcor SpO

2

cable License

9000-10-05161 - Reusable Adult Nellcor SpO

2 sensor

115-027715-00 - Reusable Adult NIBP Cuff

6200-30-09688 - NIBP Hose

Nellcor SpO

2

, 3/5-lead ECG, IBP, Advanced

Arrhythmia and ST analysis

This kit has the same contents as M51AF-PA00042 with the difference between them being that this configuration contains the Advanced ST/ARR

Software

Module Part Number 115-022719-00

Accessories kit (PN 115-023314-00) which contains:

040-001235-00 - Temperature Cable

040-001416-00 - ECG cable

0012-00-1503-02 - ECG Leadwire

0010-20-42712 - Nellcor SpO

2

cable License

9000-10-05161 - Reusable Adult Nellcor SpO

2 sensor

115-027715-00 - Reusable Adult NIBP Cuff

6200-30-09688 - NIBP Hose

Benelink module

Adult BIS 2X accessory kit

Pediatric BIS 2X accessory kit

CCO/SvO

2

Module with interface cable

2-ch IBP module

Microstream CO

2

module with accessory kit

Module Part Number: 115-030471-00

Module Part Number: 6800-30-50761

0010-10-42672 - 5 BIS Sensors (Quatro)

Module Part Number: 6800-30-50761

0010-10-42673 - 5 BIS pediatric Sensors (Quatro)

Module Part Number: 115-003480-00

009-000259-00 - CCO/SVO

2

Cable

Module Part Number: 115-029851-00

Module Part Number: 6800-30-50558

Accessories kit: 6800-30-50619 which contains

0010-10-42560 - XS-04620 Oridion Filter Line set

0010-10-42577 - 8174 Oridion NIV-Line

Passport 12m/Passport 17m Operator's Manual

Part No.

6800-30-50619

6800-30-50842

6800-30-50621

6800-30-50853

6800-30-50612

6800-30-50139

Description

Microstream CO

2

Accessory Kit

Multi-Gas Module with auto ID, O

2

and

Multi-Gas accessory kit

Multi-Gas Accessory Kit

Contents

0010-10-42560 - XS-04620 Oridion Filter Line set

0010-10-42577 - 8174 Oridion NIV-Line

Module Part Number: 6800-30-50585

Accessories kit 6800-30-50621 which contains:

9200-10-10530 - Dryline water trap,

Adult/Pediatric, box of 10

9200-10-10574 - Dryline water trap, Neonatal, box of 10

9200-10-10533 - Sampling line, Adult, 2.5 m, box of

25

9200-10-10555 - Sampling line, Neonatal, 2.5 m, box of 25

9000-10-07486 - Dryline airway adapter, straight, box of 10

9000-10-07487- Dryline airway adapter, elbow, box of 10

9200-10-10530 - Dryline water trap,

Adult/Pediatric, box of 10

9200-10-10574 - Dryline water trap, Neonatal, box of 10

9200-10-10533 - Sampling line, Adult, 2.5 m, box of

25

9200-10-10555 - Sampling line, Neonatal, 2.5 m, box of 25

9000-10-07486 - Dryline airway adapter, straight, box of 10

9000-10-07487- Dryline airway adapter, elbow, box of 10

Module Part Number: PN 6800-30-50488

Accessory kit (Part Number 6800-30-50612) which accessory kit 0010-30-42678 - Adult/Pediatric reusable flow sensor

6800-20-50328 - RM Connecting Plug

Respitory Mechanics Accessory Kit

0010-30-42678 - Adult/Pediatric reusable flow sensor

6800-20-50328 - RM Connecting Plug

Sidestream CO

2

module with adult/pediatric Module Part Number: 6800-30-50137

Passport 12m/Passport 17m Operator's Manual 36-3

Part No.

6800-30-50618

6800-30-50467

115-023310-00

115-020769-00

115-021056-00

115-021055-00

Description

CO

2

accessory kit

Contents

Accessory Kit (Part Number 6800-30-50618) which contains:

9200-10-10530 - Dryline water trap,

Adult/Pediatric, box of 10

9200-10-10533 Adult Sampling Line

9000-10-07486 - Airway Adapter

M02A-10-25937 - Adult Nasal Cannula

M02A-10-25938 - Pediatric Nasal Cannula

Adult/pediatric CO

Sidestream CO

2

2

accessory kit

kit, Neonatal

9200-10-10530 - Dryline water trap,

Adult/Pediatric, box of 10

9200-10-10533 Adult Sampling Line

9000-10-07486 - Airway Adapter

M02A-10-25937 - Adult Nasal Cannula

M02A-10-25938 - Pediatric Nasal Cannula

9200-10-10574 - Dryline water trap, Neonatal, box of 10

9200-10-10555 - Sampling line, Neonatal, 2.5 m, box of 25

M02B-10-64509 same as License 6892 (Salter Labs)

/ 4200

9000-10-07486 - Airway Adapter

Masimo module (additional-including cable and adult reusable sensor)

Masimo Accessory Kit

Module Part Number: 115-020921-00

Accessories kit (115-020769-00) which contains:

115-020768-00 - Masimo SpO2 cable

100-000078-00 - Reusable Adult Masimo SpO

2 sensor

115-020768-00 - Masimo SpO2 cable

100-000078-00 - Reusable Adult Masimo SpO

2 sensor

Sidestream CO

2

module (1X) with neonatal accessory kit

Module Part Number 115-020189-00

Accessory Kit: 115-021055-00 which contains:

9200-10-10555 - Neonate Sampling Line

100-000081-00 - Dryline II water trap

M02B-10-64509 - Neonate Nasal Cannula

040-001187-00 - Airway Adapter

9200-10-10555 - Neonate Sampling Line

Sidestream CO

2

module (1X) neonatal 100-000081-00 - Dryline II water trap, neonate accessory kit M02B-10-64509 - Neonate Nasal Cannula

040-001187-00 - Airway Adapter

36-4 Passport 12m/Passport 17m Operator's Manual

Part No.

115-020919-00

115-021054-00

Description

Sidestream CO

2

module with adult/pediatric accessory kit

Contents

Module Part Number 115-020189-00

Accessory Kit: 115-021054-00 which contains:

9200-10-10533 Adult Sampling Line

100-000080-00 - Dryline II water trap, adult

M02A-10-25937 - Adult Nasal Cannula

M02A-10-25938 - Pediatric Nasal Cannula

9000-10-07486 - Airway Adapter

Sidestream CO

2

module (1X) adult/pediatric accessory kit

9200-10-10533 Adult Sampling Line

100-000080-00 - Dryline II water trap, adult

M02A-10-25937 - Adult Nasal Cannula

M02A-10-25938 - Pediatric Nasal Cannula

9000-10-07486 - Airway Adapter

Passport 12m/Passport 17m Operator's Manual 36-5

FOR YOUR NOTES

36-6 Passport 12m/Passport 17m Operator's Manual

A

Product Specifications

NOTE

For the specifications of T1

and N1

, refer to their operating manuals.

A.1

Monitor Safety Specifications

A.1.1

Classifications

The monitor is classified, according to IEC60601-1:

Main unit

MPM

IBP module

SpO

2

module

NMT module

C.O. module

BIS module

AG module

CO

2

module

RM module

ScvO

2

module

BeneLink module

SMR

CCO/SvO

2

module

I

NA

Not marked

CF(*)

BF(*)

Not marked

Not marked

Not marked

IPX 1 Not suitable

I: Class I equipment

BF: Type BF applied part. (*Defibrillator-proof protection against electric shock.)

CF: Type CF applied part. (*Defibrillator-proof protection against electric shock.)

NA: Not applicable

IPX1: Protection against vertically falling water drops.

Not suitable: Equipment not suitable for use in the presence of a flammable anesthetic gas mixture.

Continuous

Passport 12m/Passport 17m Operator's Manual A-1

A.1.2

Environmental Specifications

CAUTION

The equipment may not meet the performance specifications if stored or used outside the specified temperature and humidity ranges.

Main unit, MPM, SpO

2

module, IBP module, C.O. module, Recorder, CCO/SvO

2

module, BIS module, BeneLink module,

NMT module

Item

Temperature (ºC)

Relative humidity (noncondensing)

Barometric (mmHg)

Operating conditions

0 to 40

15% to 95%

427.5 to 805.5

Storage conditions

-20 to 60

10% to 95%

120 to 805.5

Microstream CO

2

module

Item

Temperature (ºC)

Relative humidity (noncondensing)

Barometric (mmHg)

Operating conditions

0 to 40

15% to 95%

430 to 790

Storage conditions

-20 to 60

10% to 95%

430 to 790

Sidestream CO

2

module

Item

Temperature (ºC)

Relative humidity (noncondensing)

Barometric (mmHg)

AG module

Operating conditions

5 to 40

15% to 95%

430 to 790

Storage conditions

-20 to 60

10% to 95%

430 to 790

Item

Temperature (ºC)

Relative humidity (noncondensing)

Barometric (mmHg)

RM module

Item

Temperature (ºC)

Relative humidity (noncondensing)

Barometric (mmHg)

ScvO

2

module

Item

Temperature (ºC)

Relative humidity (noncondensing)

Barometric (mmHg)

Operating conditions

10 to 40

15% to 95%

525 to 805.5

Operating conditions

5 to 40

15% to 95%

427.5 to 805.5

Operating conditions

10 to 40

15% to 75%

427.5 to 805.5

Storage conditions

-20 to 60

10% to 95%

525 to 805.5

Storage conditions

-20 to 60

10% to 95%

120 to 805.5

Storage conditions

-20 to 60

10% to 90%

120 to 805.5

A-2 Passport 12m/Passport 17m Operator's Manual

Note

The environmental specifications of unspecified parameters are the same as those of the main unit.

A.1.3

Power Supply Specifications

Line voltage 100 to 240 VAC (±10%)

Current

Frequency

Fuse

Passport 12m: 2.5 to 1.4 A

Passport 17m: 2.8 to 1.6 A

50 Hz /60 Hz (±3 Hz)

Passport 12m: Time-lag 250V T3.15A

Passport 17m: Time-lag 250V T4A

A.2

Physical Specifications

Components

Main unit (Passport 12m)

Main unit (Passport 17m)

SMR (OEM)

SMR (Mindray)

Weight Size Equipment type

<7.5 kg (16.5 pound)

<12.0 kg (26.5 pound)

297×336×187 mm (11.7×13.2×7.4 inch) Without modules, batteries,

400×370×193 mm (15.7×14.5×7.6 inch) accessories, and recorder

<1.8 kg (4.0 pound) 142×402×151 mm (5.6×15.8×5.9 inch)

With no module inserted

<2.30 kg (5.1pound) 403× 221× 145 mm (15.9×8.7×5.7inch)

<0.63 kg (1.4 pound) 136.5×80.5×102 mm (5.4×3.2×4.0 inch)

<0.26 kg (0.57 pound) 136.5×40×102 mm (5.4×1.6×4.0 inch)

With a handle and cable hooks

<0.25 kg (0.55 pound) 136.5×40×102 mm (5.4×1.6×4.0 inch)

<0.25 kg (0.55 pound) 136.5×40×102 mm (5.4×1.6×4.0 inch)

MPM

SpO

2

module

IBP module

C.O. module

Sidestream CO

2 module (2 slots) <0.48 kg (1.1 pound)

Sidestream CO

2 module (1 slot)

Microstream CO

2

module

AG module (3 slots)

AG module (2 slots)

<0.60 kg (1.3 pound)

<0.37 kg (0.8 pound)

<1.75 kg (3.8 pound)

<1.03 kg (2.27 pound)

136.5×80.5×102 mm (5.4×3.2×4.0 inch)

136.5×40×102 mm (5.4×1.6×4.0 inch)

136.5×40×102 mm (5.4×1.6×4.0 inch)

With O

2

and BIS

136.5×121×102 mm (5.4×4.8×4.0 inch)

136.5 × 80.5 × 102 (5.4×3.2×4.0 inch) modules

With O

2

module

AG module (2 slots)

BIS module

RM module

<1.15 kg (2.53 pound)

<0.25 kg (0.55 pound)

<0.27 kg (0.59 pound)

136.5 × 80.5 × 102 (5.4×3.2×4.0 inch)

136.5×40×102 mm (5.4×1.6×4.0 inch)

136.5×40×102 mm (5.4×1.6×4.0 inch)

With O

2

and BIS modules

CCO/SvO

ScvO

2

2

module

module

BeneLink module

NMT module

<0.25 kg (0.55 pound) 136.5×40×102 mm (5.4×1.6×4.0 inch)

<0.26 kg (0.57 pound) 136.5×40×102 mm (5.4×1.6×4.0 inch)

<0.35kg (0.8 pound) 136.5×40×102 mm (5.4×1.6×4.0 inch)

<0.30 kg (0.66 pound) 136.5×40×102 mm (5.4×1.6×4.0 inch)

Passport 12m/Passport 17m Operator's Manual A-3

A.3

Hardware Specifications

A.3.1

Display

Host display

Screen type

Screen Size (diagonal)

Resolution

Color TFT LCD

12"( Passport 12m); 17"( Passport 17m)

800×600 pixels(Passport 12m); 1280×1024 pixels(Passport 17m)

Secondary display

Screen type

Screen Size

Medical-grade TFT LCD

Passport 12m: 15", 17",19"

Passport 17m: 17", 19"

A.3.2

Recorder

Method

Horizontal resolution

Vertical resolution

Thermal dot array

16 dots/mm (25 mm/s paper speed)

8 dots/mm

50 mm

20 m

25 mm/s, 50 mm/s

Maximum 3

Paper width

Paper length

Paper speed

Number of waveform channels

A.3.3

Battery

Size

Weight

Number of batteries

Battery Type

Voltage

Capacity

Run time

Charge time

Shutdown delay

147.5×60.4×23.8 mm (5.8×2.4×0.9 inch)

350 g

1 or 2 (Passport 12m); 2 (Passport 17m)

Chargeable Lithium-Ion

11.1 VDC

4500 mAh

Passport 12m: 330 minutes when powered by two new fully-charged batteries

(25°C, SpO

2

sensor connected, ECG and Temp cables disconnected, Auto NIBP measurements at an intervals of 15 minutes)

Passport 17m: 120 minutes when powered by two new fully-charged batteries

(25°C, SpO

2

sensor connected, ECG and Temp cables disconnected, Auto NIBP measurements at an intervals of 15 minutes) less than 5.5 h to 90% less than 6 h to 100% at least 5 min (after a low battery alarm first occurs)

A-4 Passport 12m/Passport 17m Operator's Manual

A.3.4

LEDs

Alarm lamp

Power on LED

AC power LED

1 (two color coded: yellow and red)

1 (green)

1 (green)

Battery LED

A.3.5

Audio Indicator

1 (green)

Speaker

Wired network

USB

Give alarm tones (45 to 85 dB), key tones, QRS tones; support PITCH TONE and multi-level tone modulation; alarm tones comply with IEC60601-1-8.

A.3.6

Monitor Interface Specifications

Power 1 AC power input connector

Passport 12m: 1 RJ45 connector, 100 Base-TX, IEEE 802.3

Passport 17m: 2 RJ45 connector, 100 Base-TX, IEEE 802.3

Passport 12m: 4 connectors, USB 1.1;

Passport 17m: up to 10 connectors, USB 1.1

SMR connector

CF

Video interface

Nurse call

Equipotential Grounding Terminal

Micro-D connector

A.3.7

Outputs

Auxiliary Output

1 connector, not standard USB

50-pin CF revision 2.0 connector

1 connector, standard DVI-D

1 connector, standard BNC

1 terminal

1 connector, standard Micro-D

Standard

Meets the requirements of IEC60601-1 for short-circuit protection and leakage current

ECG Analog Output

Bandwidth

(-3dB; reference frequency: 10Hz)

QRS delay

Sensitivity

PACE rejection/enhancement

Diagnostic mode:

Monitor mode:

Surgical mode:

ST mode:

0.05 to 150 Hz

0.5 to 40 Hz

1 to 20 Hz

0.05 to 40 Hz

≤25 ms (in diagnostic mode, and non-paced)

1V/mV ±5%

Pace enhancement

Signal amplitude: Voh≥2.5V

Pulse width: 10ms±5%

Signal rising and falling time: ≤100µs

IBP Analog Output

Bandwidth (-3dB; reference frequency:1 Hz)

DC to 40 Hz

Passport 12m/Passport 17m Operator's Manual A-5

Max transmission delay

Sensitivity

Nurse Call Signal

≤30 ms

1 V/100 mmHg ±5%

Amplitude

High level: 3.5 to 5 V, providing a maximum of 10 mA output current;

Low level: < 0.5 V, receiving a maximum of 5 mA input current.

≤1 ms Rising and falling time

Defib Sync Pulse

Output impedance

Max time delay

Amplitude

≤100Ω

35 ms (R-wave peak to leading edge of pulse)

High level: 3.5 to 5 V, providing a maximum of 10 mA output current;

Low level: < 0.5 V, receiving a maximum of 5 mA input current.

100 ms ±10% Pulse width

Rising and falling time ≤1 ms

Digital video output (DVI-D connector)

Video signals Single Link TMDS

DDC signals Signals 12C compliant

Alarm output (Network connector)

Alarm delay time from the monitor to remote equipment

The alarm delay time is equal to or smaller than 2 s, measured from the time of the monitor alarm signal generation to the time of the remote equipment alarm signal generation.

A.4

Data Storage

Trends

Parameter alarms

OxyCRG events

Arrh. events

NIBP measurements

Full-disclosure waveforms

Trends: 120 hours, at 1 min resolution

Mid-length trends: 4 hours, at 5 s resolution

Minitrends: 1 hour, at 1 s resolution

100 physiological alarms and manual events and related parameter waveforms.

100 OxyCRG events

100 arrhythmia events and relate waveforms and parameters.

1000 sets

48 hours at maximum. The specific storage time depends on the waveforms stored and the number of stored waveforms.

A-6 Passport 12m/Passport 17m Operator's Manual

A.5

Wireless Network (For Passport 12m Only)

Protocol

Modulation mode

Operating frequency

Channel spacing

Wireless baud rate (data rate)

Output power (transfer power)

Operating mode

Data security

Communication with the central station

Communication with other monitors

Data integrity

Data latency

Priority

IEEE 802.11a/b/g/n

DSSS and OFDM

2.4G Frequency Bands(FCC):2.4 GHz to 2.483GHz (only use channel 1 to channel 11)

5G Frequency Bands(FCC): 5.15 GHz to 5.35GHz, 5.47GHz to 5.725GHz, 5.725GHz to

5.85GHz

IEEE 802.11a: 20MHz

IEEE 802.11b/g: 5MHz

IEEE 802.11n at 2.4 GHz: 5MHz

IEEE 802.11n at 5 GHz: 20MHz

IEEE 802.11a: 6 to 54 Mbps

IEEE 802.11b: 1 to 11 Mbps

IEEE 802.11g: 6 to 54 Mbps

IEEE 802.11n (at both 2.4GHz and 5GHz): 6.5 to 72.2 Mbps

< 20 dBm (CE requirement: detection mode – RMS);

< 30 dBm (FCC requirement: detection mode – peak power).

Infrastructure

Standards: WPA/WPA2-PSK, WPA/WPA2-Enterprise

EAP methods: PEAP-MsCHAPv2, EAP-FAST, EAP-TLS, EAP-TTLS, PEAP-GTC, PEAP-TLS,

EAP-LEAP

Encryptions: TKIP,AES

1.

The Passport 12m transmits waveforms and numerics of parameters (e.g. ECG,

Resp, Temp, SpO

2

, CO

2

, NIBP), related alarms and alarm settings, patients' information, operating mode, and and historical data, including trends and events, to the central station. The waveforms, numerics, alarms, alarm settings, patients' information, operating mode, and historical data displayed on the central station are consistent with the Passport 12m.

2.

When so configured, the central station transmits alarm settings, patients' information, and operating mode settings to the Passport 12m. The alarm settings, patients' information, and operating mode settings of the Passport

12m are consistent with the central station.

The Passport 12m transmits waveforms, numerics of parameters (e.g. ECG, Resp,

Temp, SpO

2

, CO

2

, NIBP), and related alarms to other monitors. The waveforms, numerics, and alarms displayed on other monitors are consistent with the Passport

12m.

The probability for loss of application data within wireless network shall be less than

1000ppm (0.1%).

Total delay of data transmission from the Passport 12m to the central station: ≤ 2s

Delay of the central station to configure related settings of the Passport 12m: ≤ 2s

Total delay of data transmission from Passport 12m to other patient monitors: ≤ 2s

Delay of other patient monitors to reset the alarms of the Passport 12m: ≤ 2s

QoS setting is supported. All communication data types have the same priority at default, but real time monitoring data transmission priority can be configured with a

Passport 12m/Passport 17m Operator's Manual A-7

Transmission distance

Roaming

System capacity

Resistance to wireless interference

Dynamic networking stability

Network interruption alarm

Loss of data

FCC approval higher priority than other data transmission.

With line of sight from the Passport 12m to the AP, the transmission distance shall be no less than 50 m.

Network switchover is automatically implemented when the Passport 12m moves from the coverage area of AP1 to that of AP2.

Number of Passport 12m monitors supported by a single AP: ≤ 16. Each Passport

12m can communicate with the central station and connect to two other monitors at the same time, and among them, at most two Passport 12m can transmit historical data when reconnected at the same time. The wireless functions of all

Passport 12m monitors are normal at the same time.

The wireless functions of the Passport 12m are normal when the following conditions exist simultaneously:

1.

The distance between interfering devices (including wireless devices at the frequency of 2.4GHz, cellular communication devices, microwave ovens, intercoms, cordless phones and electro-surgical units, exclude Wi-Fi) and the monitor is greater than 20 cm.

2.

Co-channel interference (CCI) on the Wi-Fi network should be no greater than

-85dBm.

3.

Adjacent–channel interference (ACI) on the Wi-Fi network should be no greater than -50dBm.

The wireless functions of the test monitor are normal when the Passport 12m is moving at the rate of no more than 3.75 m/s within a 15m non-blocking linear distance.

1.

When network interruption occurs, the Passport 12m screen shows a disconnection icon, and the system initiates the related alarms.

2.

When the network is reconnected, wireless connection recovers automatically.

When 16 Passport 12m monitors are connected to one AP, the probability for loss of data between the subject Passport 12m monitor and the BeneVision central station is less than 0.1

in a 24 hour period, with any 12 of the 16 Passport 12m monitors roaming 30 times.

SQG-WB45NBT

A-8 Passport 12m/Passport 17m Operator's Manual

A.6

Measurement Specifications

The adjustable range for alarm limits is the same as the measurement range for the parameter unless otherwise specified.

A.6.1

ECG

ECG

Standards

Lead set

ECG standard

Display sensitivity

Sweep speed

Bandwidth (-3dB)

Meet standards of EC11, EC13, EN60601-2-27/IEC60601-2-27 and IEC60601-2-25

3-lead: I, II, III

5-lead: I, II, III, aVR, aVL, aVF, V

12-lead: I, II, III, aVR, aVL, aVF, V1 to V6

AHA, IEC

1.25 mm/mV (X0.125), 2.5 mm/mV (X0.25), 5 mm/mV (X0.5), 10 mm/mV (X1), 20 mm/mV (X2), 40 mm/mV (X4), Auto

6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s

Diagnostic mode:

Monitor mode:

Surgical mode:

ST mode:

0.05 to 150 Hz

0.5 to 40 Hz

1 to 20 Hz

0.05 to 40 Hz

Common mode rejection ratio

(with Notch off)

Notch filter

Differential input impedance

Input signal range

Accuracy of signal reproduction

Electrode offset potential tolerance

Lead-off detection current

Input offset current

Baseline recovery time

Patient leakage current

Calibration signal

ESU protection

Sampling rate

Accuracy

Diagnostic mode:

Monitor mode:

Surgical mode:

>90 dB

>105 dB

>105 dB

ST mode: >105 dB

50/60 Hz

Monitor, ST and surgical mode: Notch turns on automatically. Diagnostic mode:

Notch is turned on/off manually

≥5 MΩ

±8 mV (peak-to-peak value)

Use A and E methods based on IEC 60601-2-25 to determine frequency response.

±500 mV

Measuring electrode: <0.1 μA

Drive electrode: <1 μA

Measuring electrode: ≤0.1 μA

Drive electrode: ≤1 μA

<5 s (after defibrillation)

<10 uA

1mV (peak-to-peak value)

Cut mode: 300 W

Coagulate mode: 100 W

Recovery time: ≤10 s

In compliance with the requirements in IEC 60601-2-27: Clause 202.6.2.101

500 samples/s (A/D)

500 samples/s (ECG algorithm)

2.44μV/LSB

Passport 12m/Passport 17m Operator's Manual A-9

Pace Pulse

Pace pulse markers

Pace pulse rejection

Pace pulses meeting the following conditions are labeled with a PACE marker:

Amplitude:

Width:

Rise time:

±2 to ±700 mV

0.1 to 2 ms

10 to 100 µs

When tested in accordance with the IEC60601-2-27: 201.12.1.101.13, the heart rate meter rejects all pulses meeting the following conditions.

Amplitude:

Width:

Rise time:

±2 to ±700 mV

0.1 to 2 ms

10 to 100 µs

2.9 V/s RTI when measured in accordance with IEC60601-2-27 Section 4.1.4.3.

Pacer pulse detector rejection of fast

ECG signals

Advanced ECG algorithm

HR

Measurement range

Resolution

Accuracy

Sensitivity

Minimum QRS detection threshold (for

Advanced ECG algorithm only)

HR averaging method

Response to irregular rhythm

Response time to heart rate change

Time to alarm for tachycardia

3-, 5-, and 12-lead ECG

Neonate:

Pediatric:

Adult:

1 bpm

±1 bpm or ±1%, whichever is greater.

200μV (lead II)

Adult and pediatric: 0.16 – 0.48 mV

15 to 350 bpm

15 to 350 bpm

15 to 300 bpm

Neonate: 0.12 – 0.40 mV

In compliance with the requirements in IEC60601-2-27: Clause 201.7.9.2.9.101 b)

3) , the following method is used:

If the last 3 consecutive RR intervals are greater than 1200 ms, the 4 most recent RR intervals are averaged to compute the HR. Otherwise, heart rate is computed by subtracting the maximum and minimum ones from the most recent 12 RR intervals and then averaging them.

The HR value displayed on the monitor screen is updated every second.

In compliance with the requirements in : IEC60601-2-27: Clause 201.7.9.2.9.101 b)

4), the heart rate after 20 seconds of stabilization is displayed as follows:

Ventricular bigeminy (3a): -80±1 bpm

Slow alternating ventricular bigeminy (3b): -60±1 bpm

Rapid alternating ventricular bigeminy (3c): -120±1 bpm

Bidirectional systoles (3d): -90±2 bpm

Meets the requirements of IEC60601-2-27: Clause 201.7.9.2.9.101 b) 5).

From 80 to 120 bpm: less than 11 s

From 80 to 40 bpm: less than 11 s

Meets the requirements of IEC60601-2-27: Clause 201.7.9.2.9.101 b) 6).

A-10 Passport 12m/Passport 17m Operator's Manual

Tall T-wave rejection capability

Arrhythmia alarms

ST Segment Analysis

Measurement range

Accuracy

QT/QTc Interval Monitoring

Measurement range

Accuracy

Resolution

Alarm limit

HR High

HR Low

ST High

ST Low

QT/QTc alarm limit

QTc High

Δ

QTc High

Waveform

4ah - range: < 11 s

4a - range: < 11 s

4ad - range: < 11 s

4bh - range: < 11 s

4b - range: < 11 s

4bd - range: < 11 s

When the test is performed based on IEC60601-2-27: Clause 201.12.1.101.17 , the maximum T-wave amplitude that can be rejected is 1.2mV.

Asystole, VFib/VTac, Vtac, Vent. Brady, Extreme Tachy, Extreme Brady, PVCs/min,

Couplet, Bigeminy, Trigeminy, R on T, Run PVCs, PVC, Tachy, Brady, Missed Beats,

Vent. Rhythm, Pacer Not Pacing, Pacer Not Capture, Multif. PVC, Nonsus. Vtac,

Pause, Irr. Rhythm, AFib

-2.0 to 2.0 mV RTI

-0.8 to 0.8 mV:

Beyond this range:

±0.02 mV or ±10%, whichever is greater.

Not specified.

QT: 200 – 800 ms

QTc: 200 – 800 ms

QT-HR: 15 – 150 bpm for adult, 15 – 180 bpm for pediatric and neonate

QT: ± 30 ms

QT: 4 ms

QTc: 1 ms

Range

(low limit + 2) to 300 bpm

15 to (high limit – 2) bpm

Step

HR

40: 1bpm

HR>40: 5 bpm

(low limit +0.2) to 2.0 mV

-2.0 to (high limit – 0.2) mV

Range (bpm)

200 – 800 ms

30 – 200 ms

0.1mV

Step (ms)

10

Passport 12m/Passport 17m Operator's Manual A-11

A.6.2

Resp

Technique

Lead

Respiration excitation waveform

Respiration impedance range

Baseline impedance range

Differential input impedance

Bandwidth

Sweep speed

Respiration Rate

Measurement range

Resolution

Accuracy

Apnea alarm time

Alarm limit

RR High

RR Low

A.6.3

SpO

2

Alarm limit

SpO

2

High

SpO

2

Low

Desat

Trans-thoracic impedance

Options are lead I and II.

<300 µA RMS, ,62.8 kHz (±10%)

0.3 to 5Ω

200 to 2500Ω (using an ECG cable with 1kΩ resistance)

>2.5 MΩ

0.2 to 2.5 Hz (-3 dB)

3mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s or 50 mm/s

Adult:

Pediatric, neonate:

1 rpm

7 to 150 rpm:

0 to 6 rpm:

0 to 120 rpm

0 to 150 rpm

±2 rpm or ±2%, whichever is greater

Not specified.

10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s

Range (rpm)

Adult, pediatric:

Neonate:

0 to (high limit – 2)

(low limit + 2) to 100

(low limit + 2) to 150

Step (rpm)

1

Range (%)

(low limit + 2) to 100

Step (%)

Masimo: Desat to (high limit – 2)

Nellcor: Desat or 20 (whichever is greater) to (high limit – 2)

1

0 to (high limit – 2)

A-12 Passport 12m/Passport 17m Operator's Manual

Masimo SpO

2

Module

SpO

2

Standards

Measurement range

Resolution

Response time

Accuracy 1

Meet ISO9919 standards

1 to 100%

1%

≤ 20 s (PR 75 bpm, average time 8 s, SpO

2

value rises from 60% to 95%)

70 to 100%: ±2% (measured without motion in adult/pediatric mode)

70 to 100%: ±3% (measured without motion in neonate mode)

70 to 100%: ±3% (measured with motion)

Refreshing rate

SpO

2

averaging time

1% to 69%: Not specified.

≤ 2 s

2-4 s, 4-6 s, 8 s, 10 s, 12 s, 14 s, 16 s

Pulse amplitude: >0.02%

Low perfusion conditions

Low perfusion SpO

2

accuracy 2

PI measurement range

Light penetration: >5%

±2%

0.02 to 20%

1 The Masimo pulse oximeter with sensors have been validated for no motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies in the range of 70% to 100% SpO

2

against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population. One percent was added to the accuracies for neonatal sensors to account for accuracy variation due to properties of fetal hemoglobin.

The Masimo pulse oximeter with sensors has been validated for motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies while performing rubbing and tapping motions at 2 to 4 Hz. At an amplitude of 1 to 2 cm and non-repetitive motion between 1 to 5 Hz. At an amplitude of 2 to 3 cm in induced hypoxia studies in the range of

70% to 100% SpO

2

against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.

2 The Masimo pulse oximeter has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and Masimo's simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.

Nellcor SpO

2

Module

Measurement range

Resolution

Response time

Refreshing rate

Accuracy

0 to 100%

1%

≤ 30 s (PI > 0.3, no disturbance, SpO

2

value sudden change within 70% - 100%)

≤ 2 s

70 to 100%: ±2% (adult/pediatric)

70 to 100%: ±3% (neonate)

0% to 69%: Not specified.

When the SpO

2 sensor is applied for neonatal patients as indicated, the specified accuracy range is increased by ±1%, to compensate for the theoretical effect on oximeter measurements of fetal hemoglobin in neonatal blood.

Passport 12m/Passport 17m Operator's Manual A-13

Information of the Test Subjects of the Clinical Study Report:

Skin color

Black

Yellow

A.6.4

PR

Alarm limit

PR High

PR Low

PR from Masimo SpO

2

Module

Gender

Male

Female

Male

Female

1

3

Number

1

9

Range (bpm)

(low limit +2) to 300

15 to (high limit-2)

Measurement range

Resolution

Response time

Accuracy

Age (years)

28.2

±

9.19

Step (bpm)

1

Health

Healthy

25 to 240 bpm

1 bpm

≤ 30 s (PI > 0.3, no disturbance, PR value sudden change within 25 – 240 bpm)

±3 bpm (measured without motion)

±5 bpm (measured with motion)

PR from Nellcor SpO

2

Module

Measurement range

Resolution

Response time

Accuracy

20 to 300 bpm

1 bpm

≤ 30 s (PI > 0.3, no disturbance, PR value sudden change within 25 – 250 bpm)

20 to 250 bpm: ±3 bpm

251 to 300 bpm, not specified

PR from IBP Module

Measurement range

Resolution

Accuracy

25 to 350 bpm

1 bpm

±1 bpm or ±1%, whichever is greater

A-14 Passport 12m/Passport 17m Operator's Manual

A.6.5

NIBP

Standards

Technique

Mode of operation

Auto mode repetition intervals

STAT mode cycle time

Max measurement time

Measurement ranges

(mmHg)

Meet standards of IEC 80601-2-30

Oscillometry

Manual, Auto and STAT

1, 2, 2.5, 3, 5, 10, 15, 20, 30, 60, 90, 120, 180, 240 or 480 min

5 min

Adult, pediatric:

Neonate:

180 s

90 s

Systolic:

Diastolic:

Adult

25 to 290

10 to 250

Pediatric

25 to 240

10 to 200

15 to 215

Accuracy

Resolution

Mean: 15 to 260

Max mean error:

±

5 mmHg

Max standard deviation: 8 mmHg

1mmHg

Initial cuff inflation pressure range

(mmHg)

Default initial cuff inflation pressure

(mmHg)

Adult:

Pediatric:

Neonate:

Adult:

Pediatric:

Neonate:

Adult:

80 to 280

80 to 210

60 to 140

160

140

90

297±3 mmHg

Software overpressure protection Pediatric:

Neonate:

297±3 mmHg

147±3 mmHg

Static pressure measurement range 0 mmHg to 300 mmHg

Static pressure measurement accuracy ±3 mmHg

PR

Measurement range

Resolution

Accuracy

Alarm limit

Neonate

25 to 140

10 to 115

15 to 125

Step (mmHg)

Sys High

Sys Low

Mean High

Mean Low

Dia High

Dia Low

30 to 300 bpm

1 bpm

±3bpm or ±3%, whichever is greater

Range (mmHg)

Adult: (low limit+5) to 290

Pediatric: (low limit+5) to 240

Neonate: (low limit+5) to 140

25 to (high limit-5)

Adult: (low limit+5) to 260

Pediatric: (low limit+5) to 215

Neonate: (low limit+5) to 125

15 to (high limit-5)

Adult: (low limit+5) to 250

Pediatric: (low limit+5) to 200

Neonate: (low limit+5) to 115

10 to (high limit-5)

5

Passport 12m/Passport 17m Operator's Manual A-15

Measurement accuracy verification: In adult and pediatric modes, the blood pressure measurements measured with this device are in compliance with the Standard for Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and stardard deviation by comparing with intra-arterial or auscultatory measurements (depending on the configuration) in a typical patient population. For auscultatory reference, the 5th Korotkoff sound was used to determine the diastolic pressure.

In neonatal mode, the blood pressure measurements measured with this device are in compliance with the American

National Standard for Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and stardard deviation by comparing with intra-arterial measurements (depending on the configuration) in a typical patient population.

A.6.6

Temp

Standards

Technique

Operating mode

Measurement range

Resolution

Accuracy

Meet standard of ISO 80601-2-56

Thermal resistance

Direct mode

0 to 50 °C (32 to 122°F)

0.1 °C

±0.1 °C or ±0.2 °F (without probe)

Minimum time for accurate measurement

Minimum time between measurements

Alarm limit

T1/T2 High

T1/T2 Low

TD High

Body surface: <100 s

Body cavity: <80 s

Body surface probe: <100 s

Body cavity probe: <80 s

Range

(low limit +1) to 50 °C

(low limit +1.8) to 122 °F

0.1 to (high limit -1) °C

32 to (high limit -1.8) °F

0.1 to 50°C

0 to 90°F

Step

0.1 °C

0.1°F

A.6.7

IBP

Standards

Technique

Meet standard of EN60601-2-34/IEC60601-2-34.

Direct invasive measurement

IBP

Measurement range

Resolution

Accuracy

PPV

Measurement range

Pressure transducer

Excitement voltage

Sensitivity

Zero adjustment range

Impedance range

Volume displacement (ABBOTT)

A-16

-50 to 300 mmHg

1 mmHg

±2% or ±1 mmHg, whichever is greater (without sensor)

0% ~ 50%

5 VDC, ±2%

5 μV/V/mmHg

±

200 mmHg

300 to 3000Ω

<0.04 mm 3 /100 mmHg

Passport 12m/Passport 17m Operator's Manual

Alarm limit

Sys High

Range (mmHg) Step (mmHg)

Mean High

Dia High

(low limit + 2) to 300

1

Sys Low

Mean Low -50 to (high limit – 2)

Dia Low

A.6.8

C.O.

Measurement method

Measurement range

Resolution

Accuracy

Repeatability

Alarm range

Thermodilution method

C.O.: 0.1 to 20 L/min

TB: 23 to 43 °C

TI:

C.O.:

TB, TI:

C.O.:

TB, TI:

C.O.:

TB:

0 to 27 °C

0.1 L/min

0.1 °C

±5% or ±0.1 L /min, whichever is greater

±0.1 °C (without sensor)

±2% or ±0.1 L/min, whichever is greater

23 to 43 °C

Alarm limit Range Step

TB High

TB Low

(low limit + 1) to 43 °C

(low limit + 1.8) to 109.4°F

23 to (high limit - 1) °C

73.4 to (high limit - 1.8) °F

0.1 °C

0.1 °F

A.6.9

CCO

Operating mode

Measured parameter

Parameter alarm

Interfaces with Edwards Vigilance II®monitor, Vigileo TM monitor, or EV1000 monitor

Consistent with CCO-related parameters output by Vigilance II® monitor, Vigileo TM monitor, or EV1000 monitor

Vigilance II®: CCO/CCI, EDV/EDVI,SVR/SVRI,SV/SVI,RVEF

Vigileo

TM:

: CCO/CCI, SV/SVI, SVV, SVR/SVRI

EV1000: CCO/CCI, SVR/SVRI, SV/SVI, SVV

Signal Outputs for Vigilance II Monitor

Meets the requirements of EN 60601-1 for short-circuit protection and leakage

Standard current

Output impedance

Isolation voltage

≤ 1000Ω

1500 VAC

ECG Analog Output

Passport 12m/Passport 17m Operator's Manual A-17

Bandwidth (-3dB; reference frequency:

10Hz)

Delay (QRS complex delay)

Sensitivity

Baseline

MAP Analog Signal Output

Output voltage

Output voltage error

ST mode: 0.05~40Hz

Diagnostic mode: 0.05~150Hz

Monitor mode: 0.5~40Hz

Surgical mode: 1~20Hz

≤200ms

2V/mV (±5%)

5V±15mV

0 to 5V (0 to 500mmHg)

±5%

CVP Analog Signal Output

Output voltage

Output voltage error

Signal Outputs for Vigileo

TM

Monitor

0 to 5V (0 to 100mmHg)

±5%

Standard

Output impedance

Isolation voltage

CVP Analog Signal Output

Meets the requirements of EN 60601-1 for short-circuit protection and leakage current

≤ 1000Ω

1500 VAC

Output voltage

Output voltage error

A.6.10

SvO

2

Operating mode

0 to 5V (0 to 100mmHg)

±5%

Measured parameter

Parameter alarm

Signal Outputs for Vigilance II Monitor

Interfaces with Edwards Vigilance II®monitor, Vigileo TM monitor, or EV1000 monitor

Consistent with CCO-related parameters output by Vigilance II®monitor, Vigileo

TM monitor, or EV1000 monitor

Vigilance II®monitor, Vigileo TM monitor, and EV1000 monitor:: SvO

2

, ScvO

2

Standard

Output impedance

Isolation voltage

Meets the requirements of EN 60601-1 for short-circuit protection and leakage current

≤ 1000Ω

1500 VAC

SpO

2

Analog Signal Output

Output voltage

Output voltage error

A.6.11

ScvO

2

Measured parameters

0 to 10V (0 to 100%)

±5%

Measurement range

ScvO

2

0 to 99%

Measurement accuracy

50% to 80%:

±

3%

Other ranges: Not specified.

A-18 Passport 12m/Passport 17m Operator's Manual

A.6.12

CO

2

Measurement mode

Technique

Apnea delay time

Alarm limit

EtCO

2

High

EtCO

2

Low

FiCO

2

High awRR High

Sidestream, microstream

Infrared absorption

10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s

Range

(low limit + 2) to 99 mmHg

1 to (high limit - 2)mmHg

1 to 99 mmHg

Adult, pediatric:

Neonate:

0 to (high limit - 2) rpm

(low limit + 2) to 100 rpm

(low limit + 2) to 150 rpm

Step

1 mmHg

1 rpm awRR Low

Sidestream CO

2

Module

Standard

CO

2

Measurement range

Accuracy*

Accuracy drift

Resolution

Sample flowrate

Sample flowrate tolerance

Warm-up time

Response time

Gas sampling delay time awRR measurement range awRR measurement precision

Meet standard of ISO 80601-2-55

0 to 99 mmHg

0 to 40 mmHg:

41 to 76 mmHg:

±2 mmHg

±5% of the reading

77 to 99 mmHg: ±10% of the reading

Meet the requirement for measurement accuracy within 6 hours

1 mmHg

Adult: 70 ml/min, 100 ml/min, 120 ml/min, 150 ml/min

Pediatric, neonate: 70 ml/min, 100 ml/min

15% or 15 ml/min, whichever is greater.

1 min

Measured with a neonatal watertrap and a 2.5-meter neonatal sampling line:

<3.5 s @ 100 ml/min

<4 s @ 70 ml/min

Measured with a adult watertrap and a 2.5-meter adult sampling line:

<4.5 s @ 150 ml/min

<5.5 s @ 120 ml/min

<5.5 s @ 100 ml/min

<7 s @ 70 ml/min

Measured with a neonatal watertrap and a 2.5-meter neonatal sampling line:

<3 s @ 100 ml/min

<3.5 s @ 70 ml/min

Measured with a adult watertrap and a 2.5-meter adult sampling line:

<4 s @ 150 ml/min

<5 s @ 120 ml/min

<5 s @ 100 ml/min

<6.5 s @ 70 ml/min

0 to 120 rpm

±2 rpm

Passport 12m/Passport 17m Operator's Manual A-19

Effect of interference gases on CO

2

measurements

Gas Concentration (%)

N

2

O

Hal

Sev

Iso

Enf

Des

≤60

≤4

≤5

≤5

≤5

≤15

Quantitive effect*

±1 mmHg

±2 mmHg

*: means an extra error should be added in case of gas interference when CO

2

measurements are performed between

0-40mmHg.

Inaccuracy specifications are affected by the breath rate and I:E change. The end-tidal gas reading is within specification for breath rate below 15BPM and I:E ratio smaller than 1:1 relative to the gas readings without breath.

Microstream CO

2

Module

Standard

CO

2

Measurement range

Accuracy*

Meet standard of ISO 80601-2-55

0 to 99 mmHg

0 to 38 mmHg:

39 to 99 mmHg:

±2 mmHg

±5% of the reading+0.08% of (the reading-38)

Meet the requirement for measurement accuracy within 6 hours Accuracy drift

*This accuracy is applied to respiration rate no greater than 80 rpm. For respiration rate greater than 80 rpm and EtCO2 value greater than 18mmHg, the accuracy is 4 mmHg or ±12% of the reading, whichever is greater. For respiration rate greater than

60 rpm, the above accuracy can be achieved by using the CapnoLine H Set for Infant/Neonatal. In the presence of interfering gases, the accuracy specification deteriorates by 4% of the above accuracy.

Resolution 1 mmHg

Sample flow rate

Initialization time

Response time awRR measurement range awRR measurement accuracy

50 −

+

7.5

15

ml/min

180 s (maximum)

2.9 s (typical)

(The response time is the sum of the rise time and the delay time when using a

FilterLine of standard length)

Rise time: <190 ms (10% to 90%)

Delay time: 2.7 s (typical)

0 to 150 rpm

0 to 70 rpm:

71 to 120 rpm:

121 to 150 rpm:

±1 rpm

±2 rpm

±3 rpm

A-20 Passport 12m/Passport 17m Operator's Manual

A.6.13

tcGas

Operating mode

Parameters

tcpCO

2 tcpO

2

Interfaces with TCM CombiM or TCM TOSCA monitor

Measurement range

5 to 200 mmHg (0.7 to 26.7 kPa)

0 to 800 mmHg (0.0 to 99.9 kPa)

0 to 100 %

25 bpm to 240 bpm

0 to 1000 mW

Measurement accuracy

TOSCA Sensor 92, tc Sensor 54:

1 % CO

2

: better than 1 mmHg (0.13 kPa)

10 % CO

2

: better than 1 mmHg (0.13 kPa)

33 % CO

2

: better than 3 mmHg (0.4 kPa) tc Sensor 84:

1 % CO

2

: better than 1 mmHg (0.13 kPa)

10 % CO

2

: better than 1 mmHg (0.13 kPa)

33 % CO

2

: better than 5 mmHg (0.67 kPa) tc Sensor 84:

0 % O

2

: better than 1 mmHg (0.13 kPa)

21 % O

2

: better than 3 mmHg (0.4 kPa)

50 % O

2

: better than 5 mmHg (0.67 kPa)

90 % O

2

: better than 25 mmHg (3.33 kPa)

70 % to 100 %:

±

3 %

±

3 bpm

±

20 % of reading

SpO

2

PR

Power

A.6.14

AG

Standards

Technique

Warm-up time

Sample flow rate

Measurement range

Resolution

Iso accuracy

Full accuracy

Meet standard of ISO 80601-2-55

Infrared absorption

Iso accuracy mode:

Full accuracy mode:

Adult, pediatric:

Neonate:

Accuracy:

CO

2

:

O

2

:

N

2

O:

Des:

Sev:

Enf:

Iso:

Hal: awRR:

CO

2

: awRR:

CO

2

:

N

2

O:

Other anesthetic gases:

Gases

≤ 45 s

≤ 10 min

120, 150, 200 ml/min

70, 90, 120 ml/min

±10 ml/min or ±10%, whichever is greater

0 to 30%

0 to 100%

0 to 100%

0 to 30%

0 to 30%

0 to 30%

0 to 30%

0 to 30%

2 to 100 rpm

1 mmHg

1 rpm

±0.3%

ABS

±(8%

REL

+2%

ABS

)

8%

REL

Range (%

REL

) 1 Accuracy (%

ABS

)

Passport 12m/Passport 17m Operator's Manual A-21

Accuracy drift

Apnea alarm time

Rise time

(10 % ~ 90%)

Delay time

A-22

CO

N

O

2

2

2

O

0 to 1

1 to 5

5 to 7

7 to 10

>10

0 to 20

20 to 100

0 to 25

25 to 80

80 to 100

±1

±2

±3

±0.1

±0.2

±0.3

±0.5

Not specified

±2

±3

Des

Sev

Enf, Iso, Hal awRR

0 to 1

1 to 5

5 to 10

10 to 15

15 to 18

>18

0 to 1

1 to 5

5 to 8

>8

0 to 1

1 to 5

>5

2 to 60 rpm

>60 rpm

±0.15

±0.2

±0.4

±0.6

±1

Not specified

±0.15

±0.2

±0.4

Not specified

±0.15

±0.2

Not specified

±1 rpm

Not specified

Note 1 : The highest GAS LEVEL for a single halogenated anaesthetic gas in a gas mixture that is concealed when the anaesthetic concentration falls is 0.15/0.3% (Full/ISO accuracy).

Meet the requirement for measurement accuracy within 6 hours

10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s gas sample flow rate 120ml/min, using the DRYLINE™ or DRYLINE II watertrap and neonatal

DRYLINE™ sampling line (2.5m):

CO

N

2

2

O

≤250 ms (fall time ≤ 200ms)

≤250 ms

O

2

Hal, Iso, Sev, Des

≤600 ms

≤300 ms

Enf ≤350 ms gas sample flow rate 200ml/min, using the DRYLINE™ or DRYLINE II water trap and adult

DRYLINE™ sampling line (2.5m):

CO

2

N

2

O

≤250 ms (fall time ≤ 200 ms)

≤250 ms

O

2

Hal, Iso, Sev, Des

Enf

<4 s

≤500 ms

≤300 ms

≤350 ms

Passport 12m/Passport 17m Operator's Manual

Response time

Measured with a neonatal watertrap and a 2.5-meter neonatal sampling line:

120 ml/min:

CO

2

: ≤4s

N

2

O: ≤4.2s

O

2

: ≤4s

HAL/ISO

/

SEV

/

DES

/

ENF: ≤4.4s

Anesthetic agent limit

Primary anesthetic agent

In full accuracy mode: 0.15%

In ISO accuracy mode: 0.4%

Second anesthetic agent:

In full accuracy mode: 0.3% or 5% REL (10%

REL

for Isoflurane) of primary agent if primary agent is greater than 10%

In ISO accuracy mode: 0.5%

Inaccuracy specifications are affected by the breath rate and I:E change. The end-tidal gas reading is within specification for breath rate below 15BPM and I:E ratio smaller than 1:1 relative to the gas readings without breath; Add ±6%

REL

to inaccuracy for HAL and O

2

for breath rate larger than 15 BPM; Add ±6%

REL

to inaccuracy for all gases for breath rate larger than 30 BPM

(inaccuracy for HAL and O

2

are unspecified in this case); inaccuracy is unspecified for breath rate larger than 60 BPM.

Effect of interference gases on AG measurements

Gas

CO

2

O

2

N

2

O

Agent 1) 2)

Xenon

Helium

Concentration(%)

/

/

/

/

<100%

<50%

Quantitive effect(%

ABS

) 3)

CO

/

0.1

0.1

0.1

0.1

0.1

2

N

2

O

0

0

0.1

0.1

/

0.1

5)

Agent

1)

0

0.1

0.1

0.1

4)

0

0

Ethanol

Acetone

Methane

Saturated Isopropanol vapour

<0.1%

<1%

<1%

/

0

0.1

0.1

0.1

Unspecified

0

0.1

0.1

0

Unspecified

0

0

0

0

Unspecified

0.5

0.5

0.5

0.5

0.5 Metered dose inhaler propellants, /

1) Agent represents one of Des, Iso, Enf, Sev, and Hal.

2) Multiple agent interference on CO

2

, N

2

O and O

2

is typically the same as single agent interference.

3) For CO

2

, N

2

O and Agents, maximum interference from each gas at concentrations within specified accuracy ranges for each gas. The total interference of all gases is never larger than 5%

REL

.

4) Applicable to type A AG module, representing the interference effect of secondary anesthetic agents on primary anesthetic agent.

5) Measurement interference to type M AG module originates from the applied anesthetic agent that is configured manually.

Alarm limit Range Step

O

2

0.

2

/

0.2

1

0.5

0.5

EtCO

2

High (low limit + 2) to 99 mmHg

EtCO

2

Low

FiCO

2

High awRR High

1 to (high limit - 2)mmHg

1 to 99 mmHg

Adult, pediatric:

Neonate:

(low limit + 2) to 100 rpm

(low limit + 2) to 150 rpm

1 mmHg

1 rpm

Passport 12m/Passport 17m Operator's Manual A-23

awRR Low

EtO

2

High

EtO

2

Low

FiO

2

High

FiO

2

Low

EtN

2

O High

EtN

2

O Low

FiN

2

O High

FiN

2

O Low

EtHal/Enf/Iso High

EtHal/Enf/Iso Low

FiHal/Enf/Iso High

FiHal/Enf/Iso Low

EtSev High

EtSev Low

FiSev High

FiSev Low

EtDes High

EtDes Low

FiDes High

FiDes Low

A.6.15

BIS

Standards

Technique

Measured parameters

0 to (high limit - 2)rpm

(low limit + 2) to 100 %

0% to (high limit - 2)%

(low limit + 2) to 100 %

18% to (high limit - 2)%

(low limit + 2) to 100 %

0 to (high limit - 2)%

(low limit + 2) to 100 %

0 to (high limit - 2)%

(low limit + 0.2) to 5.0 %

0 to (high limit - 0.2)%

(low limit + 0.2) to 5.0 %

0 to (high limit - 0.2)%

(low limit + 0.2) to 8.0 %

0 to (high limit - 0.2)%

(low limit + 0.2) to 8.0 %

0 to (high limit - 0.2)%

(low limit + 0.2) to 18.0 %

0 to (high limit - 0.2)%

(low limit + 0.2) to 18.0 %

0 to (high limit - 0.2)%

1%

1%

0.1%

0.1%

0.1%

Calculated parameters

Impedance range

Sweep speed

Input impedance

A-24

Meet standard of IEC 60601-2-26

Bispectral index

EEG

BIS, BIS L, BIS R: 0 to 100

SQI, SQI L, SQI R: 0 to 100

EMG, EMG L, EMG R: 0 to 100 dB

SR, SR L, SR R: 0 to 100%

SEF, SEF L, SEF R: 0.5 to 30.0 Hz

TP, TP L, TP R: 40 to 100 dB

BC, BC L, BC R: 0 to 30 sBIS L, sBIS R: 0 to 10.0 sEMG L, sEMG R: 0 to 10.0

ASYM: 0 to 100%

0 to 999 kΩ

6.25 mm/s, 12.5 mm/s, 25 mm/s or 50 mm/s

>5 MΩ

Passport 12m/Passport 17m Operator's Manual

Noise (RTI)

Input signal range

EEG bandwidth

Patient leakage current

Alarm limit

BIS High

<0.3 µV (0.25 to 50 Hz)

±1 mV

0.25 to 100 Hz

<10 µA

Range

(low limit + 2) to 100

BIS Low

A.6.16

NMT

NMT from Mindray NMT module

0 to (high limit – 2)

Stimulation output

ST mode

TOF mode

PTC mode

DBS mode

NMT Block Recovery threshold

Step

1

Pulse width

100, 200, or 300 μs; monophasic rectangle pulse

Accuracy: ±10%

Current range

0 - 60 mA in increments of 5 mA

Accuracy: ± 5% or ± 2 mA, whichever is greater

Max. skin impedance 3 kΩ @ 60 mA, 5 kΩ @ 40 mA

Max. output voltage 300 V

ST-Ratio 0 - 200%

Measurement interval

TOF-Count

Manual, 1 s, 10 s, 20 s

0 - 4

TOF-Ratio

Measurement interval

PTC

Measurement interval

5 - 160%

Manual, 12s, 15s, 20s, 30s, 1min, 5min, 15min, 30min,

60min

0 - 20

Manual

Measurement interval

DBS-Count

Manual, 15s, 20s, 30s, 1min, 5min, 15min, 30min, 60min

0 - 2

DBS-Ratio 5 - 160%

Off, 1, 2, 3, 4, 5%, 10%, 20%, 30%, 40%, 50%, 60%, 70%, 80%, 90%, 100%

Passport 12m/Passport 17m Operator's Manual A-25

NMT from TOF-Watch® SX monitor

Operating mode

Parameters

TOF-Ratio

TOF-Count

Single

PTC

Interfaces with TOF-Watch® SX monitor

Measurement range

1% to 160%

0 to 4

0% to 160%

0 to 15

20.0 ºC to 41.5 ºC Tskin

A.6.17

RM

Technique

Frequency response

Dead space

Flow sensor

≥30 Hz

≤11 ml

Flow

Measurement range

Accuracy

Adult/pediatric*:

Infant:

Adult/pediatric*:

Infant:

0.1 L/min

± (2 to 120) L/min

± (0.5 to 30) L/min

1.2 L/min or ±10% of the reading, whichever is greater

0.5 L/min or ±10% of the reading, whichever is greater

Resolution

Paw

Measurement range

Accuracy

Resolution

MVe/MVi

-20 to 120 cmH

±3%×reading

0.1 cmH

2

O

2

O

Measurement range

Adult/Pediatric*:

Infant:

±10%×reading

2 to 60 L/min

0.5 to 15 L/min

Accuracy

TVe/TVi

Measurement range

Resolution

Accuracy

Adult/Pediatric*:

Infant:

1 ml

Adult/pediatric*:

Infant:

100 to 1500 ml

20 to 500 ml

±10% × reading or ± 15 ml, whichever is greater

±10% × reading or ± 6 ml, whichever is greater

RR (RM)

Measurement range

Accuracy

4 to 120 rpm

4 to 99 rpm

100 to 120 rpm

±1 rpm

±2 rpm

*Pediatric in this form does not include neonate and infant.

A-26 Passport 12m/Passport 17m Operator's Manual

TV

RSBI

Calculated Parameters

I:E

FEV1.0%

Pmean

MV

PEEP

PEF

PIF

PIP

Pplat

Compl

Alarm limit

RR High

RR Low

PEEP High

Measurement range

4:1 to 1:8

0 to 100%

0 to 120 cmH

2

20 to 1500 ml

O

0.5 to 60 L

0 to 120 cmH

2

O

2 to 120 L/min

2 to 120 L/min

0 to 120 cmH

2

O

0 to 120 cmH

2

O

0 to 200 ml/cmH

2

O

0 to 4095 rpm/L

Range

Adult, pediatric:

Neonate:

0 to (high limit -2) rpm

(low limit +1) to 120 cmH

2

O

Not specified.

(low limit + 2) to 100 rpm

(low limit + 2) to 150 rpm

Measurement accuracy

Not specified.

Not specified.

±10%×reading

Adult/pediatric: ±10% × readin or ±15 ml, whichever is greater.

Infant: ±10% × readin or ±6 ml, whichever is greater.

±10%×reading

Not specified.

±10% × reading

±10% × reading

±10% × reading

Step

1 rpm

1 cmH

2

O

PEEP Low

PIP High

0 to (high limit -1) cmH

2

O

(low limit +1) to 120 cmH

2

O

1 cmH

2

O

PIP Low

MVe High

1 to (high limit -1) cmH

2

O

(low limit +1.0) to 60.0 L/min

0.5 L/min

MVe Low 0.5 to (high limit -1.0)

Passport 12m/Passport 17m Operator's Manual A-27

FOR YOUR NOTES

A-28 Passport 12m/Passport 17m Operator's Manual

B

EMC and Radio Regulatory Compliance

B.1

EMC

The device meets the requirements of IEC 60601-1-2. All the accessories listed in Chapter 36 also meet the requirements of IEC 60601-1-2 when in use with this device.

CAUTIONS

Using accessories, transducers and cables other than those specified may result in increased electromagnetic emission or decreased electromagnetic immunity of the equipment.

The device or its components should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device or its components should be observed to verify normal operation in the configuration in which it will be used.

The device needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided below.

Other devices may interfere with this device even though they meet the requirements of CISPR.

When the inputted signal is below the minimum amplitude provided in technical specifications, erroneous measurements could result.

Portable and mobile communication equipment may affect the performance of this device.

Other devices that have RF transmitter or source may affect this device (e.g. cell phones, PDAs, and PCs with wireless function).

Guidance and Declaration - Electromagnetic Emissions

The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.

Emission tests

Radio frequency (RF) emissions CISPR 11

RF emissions CISPR 11

Harmonic emissions

IEC61000-3-2

Voltage

Fluctuations/Flicker

Emissions IEC 61000-3-3

Compliance

Group 1

Class A

Class A

Complies

Electromagnetic environment - guidance

The device uses RF energy only for its internal function.

Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

The device is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

Passport 12m/Passport 17m Operator's Manual B-1

WARNING

This equipment/system is intended for use by healthcare professionals only. This equipment/ system may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as re-orienting or relocating the [ME EQUIPMENT or ME SYSTEM] or shielding the location.

If the system is operated within the electromagnetic environment listed in

Table Guidance and declaration

electromagnetic immunity

, the system will remain safe and provide the following essential performance,

 Operating mode

Accuracy

Function

Accessories identification

Data stored

Alarm

Detect for connection

Guidance and Declaration - Electromagnetic Immunity

The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.

Immunity test IEC60601 test level Compliance level

Electromagnetic environment

- guidance

Electrostatic discharge (ESD) IEC

61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

±2 kV for power supply lines

±1 kV for input/output lines

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.

Mains power quality should be that of a typical commercial or hospital environment.

Electrical fast transient/burst IEC

61000-4-4

±2 kV for power supply lines

±1 kV for input/output lines

Surge IEC 61000-4-5 ±1 kV line(s) to line(s)

±2 kV line(s) to earth

Voltage dips, short interruptions and voltage variations on power supply input lines IEC

61000-4-11

<5 % U cycle

40 % U

T cycles

T

(>95 % dip in U

(60 % dip in U

T

T

) for 0.5

) for 5

70 % U

T

(30 % dip in U

T

) for 25 cycles

<5 % U

T

(>95 % dip in U

T

) for 5 s

±1 kV line(s) to line(s)

±2 kV line(s) to earth

<5 % U

T

(>95 % dip in U

T

) for

0.5 cycle

40 % U

T

(60 % dip in U

T

) for 5 cycles

70 % U

T

25 cycles

<5 % U

T

5 s

(30 % dip in U

T

) for

(>95 % dip in U

T

) for

Mains power quality should be that of a typical commercial or hospital environment. If the user of our product requires continued operation during power mains interruptions, it is recommended that our product be powered from an uninterruptible power supply or a battery.

B-2 Passport 12m/Passport 17m Operator's Manual

Power frequency

(50/60 HZ) magnetic field IEC 61000-4-8

3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

Note: U

T

is the AC mains voltage prior to application of the test level.

Guidance and Declaration - Electromagnetic Immunity

The device is intended for use in the specified electromagnetic environment. The customer or the user of the device should assure that it is used in such an environment as described below.

Immunity test IEC60601 test level

Compliance level Electromagnetic environment - guidance

Conduced RF

IEC61000-4-6

3 Vrms

150 kHz to 80MHz

3Vrms

(BIS, ICG: 1Vrms)

Radiated RF

IEC61000-4-3

3V/m

80MHz to 2.5GHz

3V/m

(Resp: 1V/m)

Portable and mobile RF communications equipment should be used no closer to any part of the system, including cables, than the recommended separation distance calculated from the equation appropriate for the frequency of the transmitter. Recommended separation distances:

d

= 1 .

2

For BIS, ICG:

P d

= 3 .

5

P

Recommended separation distances:

80 MHz

800 MHz d

= 1

.

2

P

800MHz-2.5GHz

(Resp:

d

= 3 .

5

P

)

d

= 2 .

3

P

(Resp:

d

= 7

P

)

Where,

P

is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and

d

is the recommended separation distance in meters (m).

b

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a , should be less than the compliance level in each frequency range b

.

Interference may occur in the vicinity of equipment marked with the following symbol:

Note 1: At 80 MHz to 800 MHz, the separation distance for the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Note 3: The device that intentionally receives RF electromagnetic energy at the exclusion band (2395.825MHz-2487.645MHz) is exempt from the essential performance requirements, but remains safe. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the [ME EQUIPMENT or ME SYSTEM] is used exceeds the applicable RF compliance level above, the [ME EQUIPMENT or ME SYSTEM] should be observed to verify normal operation. If abnormal

Passport 12m/Passport 17m Operator's Manual B-3

performance is observed, additional measures may be necessary, such as re-orienting or relocating the [ME EQUIPMENT or

ME SYSTEM]. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than3V/m. For BIS and ICG, field strengths should be less than 1V/m.

WARNING

The device is configured with a wireless network connector to receive wireless signal. Other devices may interfere with this device even though they meet the requirements of CISPR.

Recommended separation distances between portable and mobile RF communications equipment and the device

The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.

Separation distance in meters (m) according to frequency of the transmitter Rated maximum output power of transmitter (W)

150 kHz to 80 MHz

d

= 1 .

2

( BIS, ICG:

d

P

= 3 .

5

P

)

80 MHz to 800 MHz

d

= 1 .

2

(Resp:

d

=

P

3 .

5

P

)

800 MHz to 2.5 GHz

d

=

(Resp:

2 .

3

P d

= 7

P

)

0.01

0.1

0.12 (BIS, ICG: 0.35)

0.38 (BIS, ICG: 1.11)

0.12 (0.35)

0.38 (1.11)

0.23 (0.70)

0.73 (2.22)

1

10

1.20 (BIS, ICG: 3.50)

3.80 (BIS, ICG: 11.07)

1.20 (3.50)

3.80 (11.07)

2.30 (7.00)

7.30 (22.14)

100 12.00 (BIS, ICG:35.00) 12.00 (35.00) 23.00 (70.00)

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

B-4 Passport 12m/Passport 17m Operator's Manual

B.2

Radio Regulatory Compliance

RF parameters (for Passport 12m)

Type of Radio

Modulation mode

Operating frequency

IEEE 802.11b/g/n (2.4G)

DSSS and OFDM

2400-2483.5MHz

IEEE 802.11a/n (5G)

OFDM

5150-5350 MHz

5470-5725 MHz

5725-5850 MHz

20 MHz Channel spacing

Output power

5 MHz

< 30 dBm (Peak Power)

< 20 dBm (Average Power)

Federal Communications Commission (FCC) Statement

The wireless module has been tested and found to comply with the limits for a class B digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

Reorient or relocate the receiving antenna.

 Increase the separation between the equipment and receiver.

 Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.

 Consult the dealer or an experienced radio/TV technician for help.

CAUTION

Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.

The radio device used in this product is in compliance with the essential requirements and other relevant provisions of

Directive 1999/5/EC (Radio Equipment and Telecommunications Terminal Equipment Directive).

WARNING

This compliance to general radiation exposure limits for an uncontrolled environment, and minimum of 20 cm separation between monitor and human body.

Passport 12m/Passport 17m Operator's Manual B-5

FOR YOUR NOTES

B-6 Passport 12m/Passport 17m Operator's Manual

C

Default Configurations

This chapter lists some of the most important factory default settings in configuration management. You cannot change the factory default configuration itself. However, you can make changes to the settings from the factory default configuration and then save the changed configuration as a user configuration.

C.1

Parameters Configuration

C.1.1

ECG

ECG Setup

Item Name

Alm Source

Alarm

Alm Lev

Configurable

In Config Mode

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Default

HR/PR High

HR/PR Low

Gain

Sweep

Filter

QRS Volume

QRS threshold

Lead Set

ECG Display

Notch Filter

Paced

Pacer Reject

Smart Lead Off

Baseline Drift Removal

(12-lead screen only)

Yes

Yes

Yes

Yes

Yes

Yes

No

Yes

Yes

Yes

No

No

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

HR

On

Med

Adu: 120

Ped: 160

Neo: 200

Adu: 50

Ped: 75

Neo: 100

X1

25 mm/s

General: Monitor

OR: Surgery

ICU, NICU: Monitor

CCU: Diagnostic

General, OR: 2

ICU, NICU, CCU: 1

0.16 mV

Auto (if auto lead detection is available); 5-lead (if auto lead detection is not available)

Normal

Weak

No

Off

On

On

Passport 12m/Passport 17m Operator's Manual C-1

Item Name

Item Name

ST Analysis

Configurable

In Config Mode

Waveform Layout (12-lead screen only)

ECG Waveform Area

(12-lead screen only)

ST Analysis

Yes

No

Configurable

In Config Mode

Yes

In Monitor Mode

Yes

Yes

Default

Standard

Normal

ST Display

Alarm

Alm Lev

ST-X High

ST-X Low

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Yes

Auto

ISO

J

ST

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

X represents I, II, III, aVR, aVL, aVF, V, V1, V2, V3, V4, V5 or V6.

QT/QTc Analysis

Item Name

Configurable

In Config Mode

QT Analysis

QTc Formula

Analysis lead

QTc Alarm

ΔQTc Alarm

QTc Alarm Level

ΔQTc Alarm Level

QTc Record

ΔQTc Record

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Default

Off

Hodges

All

Off

Off

Med

Med

Off

Off

Default

General, OR, ICU, NICU: Off

CCU: On

All

Off

Med

When ST Unit is mV:

When ST Unit is mm:

When ST Unit is mV:

When ST Unit is mm:

Off

-80 ms

48 ms

J + 60 ms

0.20

2.0

-0.20

-2.0

C-2 Passport 12m/Passport 17m Operator's Manual

Arrh. Analysis

Arrhythmia Threshold Settings (Mindray)

Configurable

Item Name

In Config Mode

PVCs High

Asys. Delay

Vtac Rate

Vtac PVC

Multif. PVC’s Window

Yes

Yes

Yes

Yes

Yes

Tachy Yes

Brady

Extreme Tachy

Yes

Yes

Extreme Brady Yes

Vbrd Rate

Vbrd PVCs

Pause Time

Yes

Yes

Yes

Arrhythmia Alarm Settings (Mindray)

Configurable

Item Name

Asystole Alarm

In Config Mode

Yes

VFib/VTac Alarm

Vtac Alarm

Vent. Brady Alarm

Extreme Tachy Alarm

Extreme Brady Alarm

Yes

Yes

Yes

Yes

Yes

PVCs/min Alarm Yes

R on T Alarm

Run PVCs Alarm

Couplet Alarm

Multif. PVC Alarm

PVC Alarm

Bigeminy Alarm

Trigeminy Alarm

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Passport 12m/Passport 17m Operator's Manual

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Default

Yes

Yes

Yes

Yes

Yes

Yes

Yes

40

5

2

10

5

130

6

15

Adu:

Ped:

Adu:

Ped:

Adu:

Ped:

Adu:

Ped:

160

180

35

50

120

160

50

75

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Default

Alarm Switch

On

On

On

On

On

On

General, OR, ICU, NICU:

Off

CCU: On

General, OR, ICU, NICU:

Off

CCU: On

Off

Off

Off

Off

General, OR, ICU, NICU:

Off

CCU: On

General, OR, ICU, NICU:

Off

Alarm Level

High

High

High

High

High

High

Med

Med

Low

Message

Med

Message

Med

Med

C-3

Arrhythmia Threshold Settings (Mindray)

Item Name Configurable

Tachy Alarm

Brady Alarm

Pacer Not Pacing Alarm

Pacer Not Capture Alarm

Missed Beats Alarm

Nonsus. Vtac Alarm

Vent. Rhythm Alarm

Pause Alarm

Irr. Rhythm Alarm

Afib Alarm

C.1.2

RESP

Item Name

Alarm

Alm Lev

RR High

RR Low

Apnea Delay

Lead

Gain

Sweep

Detection Mode

RR Source

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Configurable

In Config Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

No

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Default

CCU: On

Off

Off

Off

Off

Off

General, OR, ICU, NICU:

Off

CCU: On

Off

Off

Off

Off

Med

Med

Message

Message

Message

Med

Med

Low

Message

Message

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Default

On

Med

Adu, Ped:

Neo:

Adu, Ped:

Neo:

Adu, Ped:

Neo:

Adu, Ped:

Neo:

X2

6.25 mm/s

Auto

Auto

30

100

8

30

20

15

Auto

II

C-4 Passport 12m/Passport 17m Operator's Manual

C.1.3

PR

Item Name

Alarm

Alm Lev

HR/PR High

Configurable

In Config Mode

Yes

Yes

Yes

HR/PR Low Yes

PR Source

Alm Source

Beat Vol

C.1.4

SpO

2

Item Name

Alarm

Alm Lev

Yes

Yes

Yes

Configurable

In Config Mode

Yes

Yes

SpO

2

High Yes

SpO

2

Low

Desat

Sensivity

Mindray

Sensivity

Masimo

Averaging

Masimo

Sat-Seconds

Nellcor

PI Zoom

Sweep

NIBP Simul

C.1.5

Δ

SpO

2

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

No

Item Name

Alarm

Alm Lev

Configurable

In Config Mode

Yes

Yes

Δ

SpO

2

High

Yes

PI Zoom Yes

Passport 12m/Passport 17m Operator's Manual

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Default

On

Med

Adu:

Ped:

Neo:

Adu:

Ped:

Neo:

SpO

2

HR

General, OR: 2

ICU, NICU, CCU: 1

120

160

200

50

75

100

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Default

On

SpO

2:

: Med

Desat: High

Adu, Ped: 100

Neo: 95

90

80

Med

Normal

8 s

0 s

No

25 mm/s

Off

In Monitor Mode

Yes

Yes

Yes

Yes

Default

Off

Med

10 %

No

C-5

C.1.6

Temp

Item Name

Alarm

Alm Lev

T1/T2 High

ºC

T1/T2 Low

ºC

TD High

ºC

C.1.7

NIBP

Item Name

Alarm

Alm Lev

Configurable

In Config Mode

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Default

On

Med

38.0

35.0

2.0

Interval

NIBP End Tone

Clock

Initial Pressure (mmHg)

Yes

Yes

Yes

Cuff Press. (mmHg)

Configurable

In Config Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Default

On

Med

General: 15 min

OR: 5 min

ICU: 15 min

NICU: 30 min

CCU: 15 min

Off

On

Adu:

Ped:

Neo:

160

140

90

Adu:

Ped:

Neo:

80

60

40

Alarm Limit

NIBP-S High (mmHg)

NIBP-S Low (mmHg)

NIBP-M High (mmHg)

Yes

Yes

Yes

Yes

Yes

Yes

Adu:

Ped:

Neo:

Adu:

Ped:

Neo:

Adu:

Ped:

Neo:

90

70

40

160

120

90

110

90

70

C-6 Passport 12m/Passport 17m Operator's Manual

Item Name

NIBP-M Low (mmHg)

NIBP-D High (mmHg)

NIBP-D Low (mmHg)

Configurable

In Config Mode

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Default

Adu:

Ped:

Neo:

Adu:

Ped:

Neo:

Adu:

Ped:

Neo:

C.1.8

IBP

Item Name

Alarm

Alm Lev

P1 Measure

P2 Measure

P3 Measure

P4 Measure

PPV Measurement

Configurable

In Config Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

PPV Source

Sensitivity

Sweep

Filter

Gridlines

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Art, Ao, UAP, BAP, FAP, LV, P1-P2 Arterial Pressure Alarm Limits

Default

On

Med

All

All

Mean

Mean

Off

Auto

Med

25 mm/s

12.5 Hz

Off

IBP-S High (mmHg)

IBP-S Low (mmHg)

IBP-M High (mmHg)

IBP-M Low (mmHg)

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Adu:

Ped:

Neo:

Adu:

Ped:

Neo:

Adu:

Ped:

Neo:

Adu:

Ped:

Neo:

Passport 12m/Passport 17m Operator's Manual

90

70

60

60

50

25

50

40

20

160

120

90

70

50

35

90

70

55

110

90

70

C-7

Item Name

IBP-D High (mmHg)

IBP-D Low (mmHg)

Configurable

In Config Mode

Yes

Yes

In Monitor Mode

Yes

Yes

Default

Adu:

Ped:

Neo:

Adu:

Ped:

Neo:

PA Alarm Limit

PA-S High (mmHg) Yes Yes

PA-S Low (mmHg)

PA-M High (mmHg)

PA-M Low (mmHg)

Yes

Yes

Yes

Yes

Yes

Yes

PA-D High (mmHg) Yes Yes

PA-D Low (mmHg) Yes Yes

CVP, LAP, RAP, ICP, UVP, P3-P4 Venous Pressure Alarm Limits

IBP-M High (mmHg) Yes Yes

IBP-M Low (mmHg) Yes Yes

Art, Ao, BAP, FAP, LV, P1-P2 Arterial Pressure Scale

Scale (mmHg) Yes Yes

PA Waveform Scale

Scale (mmHg) Yes

CVP, LAP, RAP, ICP, UVP Scale

Scale (mmHg) Yes

Yes

Yes

UAP, P3-P4 Venous Pressure Scale

Scale (mmHg) Yes

IBP Overlapping Left Scale

Scale (mmHg) Yes

IBP Overlapping Right Scale

Scale (mmHg) Yes

Yes

Yes

Yes

Adu:

Ped, Neo:

Adu:

Ped, Neo:

Adu:

Ped, Neo:

Adu:

Ped, Neo:

Adu:

Ped, Neo:

Adu:

Ped, Neo:

35

60

16

4

0

-4

10

24

20

26

0

12

Adu:

Ped, Neo:

0

10

4

0-160

0-30

0-20

0-80

0-160

0-20

50

40

20

90

70

60

C-8 Passport 12m/Passport 17m Operator's Manual

CCO Low

CCI High

CCI Low

EDV High

EDV Low

EDVI High

EDVI Low

SVR High

SVR Low

SVRI High

SVRI Low

RVEF High

RVEF Low

SV High

SV Low

SVI High

SVI Low

SvO

2

High

SvO

2

Low

ScvO

2

High

ScvO

2

Low

C.1.9

C.O.

Item Name

Alarm

Alm Lev

TB High

ºC

TB Low

ºC

Comp. Const

Auto TI

Manual TI

ºC

Measuring mode

Configurable

In Config Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

C.1.10

CCO/SvO

2

Setup (Vigilance II)

Configurable

Item Name

In Config Mode

Alarm

Alarm Level

Primary Parameter

Secondary Parameters

CCO High

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Default

On

Med

39.0

36.0

0.542

Auto

2.0

Manual

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Default

120

20

60

10

90

40

90

40

2

10

1

300

80

150

Off

Med

C.O./CCO

SVR, EDV, SV

14

60

1500 DS/cm 5

500 DS/cm 5

3000 DS·m 2 / cm 5

1000 DS·m

2

/ cm

5

50

0

Passport 12m/Passport 17m Operator's Manual C-9

Alarm Level

Primary Parameter

Secondary Parameters

CCO High

CCO Low

CCI High

CCI Low

SV High

SV Low

SVI High

SVI Low

SVV High

SVV Low

SVR High

SVR Low

SVRI High

SVRI Low

SvO

2

High

SvO

2

Low

ScvO

2

High

ScvO

2

Low

C.1.11

CCO/SvO

2

Setup (Vigileo)

Configurable

Item Name

In Config Mode

Alarm Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Default

Off

2

10

1

Med

CCO

SV, SVR, SVV

14

120

20

60

10

30

0

1500 DS/cm

5

500 DS/cm 5

3000 DS·m 2 / cm 5

1000 DS·m 2 / cm 5

90

40

90

40

C-10 Passport 12m/Passport 17m Operator's Manual

C.1.12

CCO/SvO

2

Setup (EV1000)

Configurable

Item Name

In Config Mode

Alarm Yes

Yes

Yes

Alarm Level

Parameter Display (when the

EV1000 is in the None or

VolumeView mode)

Secondary Parameters (when the EV1000 is in the None or

VolumeView mode)

Parameter Display (when the

EV1000 is in the FloTrac or

ClearSightmode)

SVI Low

SVV High

SVV Low

SVR High

SVR Low

SVRI High

SVRI Low

SvO

2

High

SvO

2

Low

ScvO

2

High

ScvO

2

Low

Secondary Parameters (when the EV1000 is in the FloTrac or

ClearSightmode)

CCO High

CCO Low

CCI High

CCI Low

SV High

SV Low

SVI High

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Default

Off

Med

Indexed

GEDI, SVRI, ELWI

Indexed

SVI, SVV, SVRI

90

40

90

40

14

2

10

1

120

20

60

10

30

0

1500 DS/cm

5

500 DS/cm

5

3000 DS·m

2

/ cm

5

1000 DS·m

2

/ cm

5

Passport 12m/Passport 17m Operator's Manual C-11

C.1.13

CO

2

Item Name

Alarm

Alm Lev

Apnea Delay

Operating Mode

Sweep

Scale (mmHg)

RR Source

Sidestream CO

2

Setup

Flow Rate

BTPS Compen

N

2

O Compen

O

2

Compen Yes

Des Compen

Microstream CO

2

Setup

BTPS Compen

Max Hold

Auto Standby (min)

Alarm Limits

EtCO

2

High (mmHg)

Yes

Yes

Yes

Yes

Yes

EtCO

2

Low (mmHg)

FiCO

2

High (mmHg)

RR High

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

No

Configurable

In Config Mode

Yes

Yes

Yes

RR Low Yes

C.1.14

tcGas

Configurable

Item Name

Alarm Sound

In Config Mode

Yes

Change Secondary Parameters Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Default

On

Med

Adu, Ped:

Neo:

Measure

6.25 mm/s

50

Auto

20

15

Adu,

Ped:

Neo:

Off

0

General: 21

OR: 100

ICU, NICU, CCU: 21

0

120 ml/min

100 ml/min

70 ml/min

Off

20 s

0

Adu, Ped:

Neo:

Adu, Ped:

Neo:

Adu, Ped, Neo: 4

50

45

25

30

Adu, Ped:

Neo:

Adu, Ped:

Neo:

30

100

8

30

Default

Off

SpO

2

, PR, Power

C-12 Passport 12m/Passport 17m Operator's Manual

C.1.15

AG

Item Name

Alarm

Alm Lev

Sweep

O

2

Compen

Operating Mode

Flow Rate

Auto Standby

Apnea Time

RR Source

CO

2

Setup

Wave Type

Scale

EtCO

2

High

( mmHg

Yes

Yes

Yes

EtCO

2

Low

( mmHg

FiCO

2

High

( mmHg

RR High

Yes

Yes

Yes

RR Low

Gas Setup

Agent

N

2

O Scale

O

2

Scale

AA Scale

Hal/Enf/Iso Scale

Des Scale

Sev Scale

EtO

2

High

EtO

2

Low

FiO

2

High

FiO

2

Low

EtN

2

O High

EtN

2

O Low

FiN

2

O High

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Passport 12m/Passport 17m Operator's Manual

Configurable

In Config Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

No

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Default

On

Med

6.25 mm/s

OR:

General, ICU, NICU, CCU:

Measure

Adu, Ped:

Neo:

Off

20 s

Auto

On

Off

120 ml/min

70 ml/min

Draw

When unit is mmHg:

When unit is % or KPa:

Adu, Ped:

Neo:

Adu, Ped:

Neo:

4

Adu, Ped:

Neo:

Adu, Ped:

Neo:

50

7.0

50

45

25

30

30

100

8

30

9.0

2.5

9.0

4.0

AA

50

When unit is mmHg:

When unit is % or KPa:

88

18

100

90

Adu, Ped:

Neo:

18

55

0

53

400

50

C-13

Item Name

FiN

2

O Low

EtHal/Enf/Iso High

EtHal/Enf/Iso Low

FiHal/Enf/Iso High

FiHal/Enf/Iso Low

EtSev High

EtSev Low

FiSev High

FiSev Low

EtDes High

EtDes Low

FiDes High

FiDes Low

C.1.16

BIS

Item Name

Alarm

Alm Lev

Smoothing Rate

Configurable

In Config Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Yes

Default

0.0

6.0

0.0

0.0

5.0

0.0

8.0

0

3.0

0.0

2.0

0.0

6.0

Display

Filters

Scale

Sweep

Trend Length

C-14

Yes

Yes

Yes

Yes

Yes

Configurable

In Config Mode

Yes

Yes

In Monitor Mode

Yes

Yes

Yes Yes

Yes

Yes

Yes

Yes

Yes

Default

On

Med

General, OR, ICU,CCU: 15 s

NICU:

Adu: 15s

Ped: 15s

Neo: N/A

General, OR, ICU,CCU: EEG

NICU:

Adu: EEG

Ped: EEG

Neo: N/A

General, OR, ICU,CCU: On

NICU:

Adu: On

Ped: On

Neo: N/A

General, OR, ICU,CCU: 100

μ

V

NICU:

Adu: 100

μ

V

Ped: 100

μ

V

Neo: N/A

General, OR, ICU,CCU: 25mm/s

NICU:

Adu: 25mm/s

Ped: 25mm/s

Neo: N/A

General, OR, ICU,CCU: 60 min

Passport 12m/Passport 17m Operator's Manual

Item Name

Secondary Parameters

Display

EEG Waveforms

Parameter 1

Configurable

In Config Mode In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Default

NICU:

Adu: 60 min

Ped: 60 min

Neo: N/A

General, OR, ICU,CCU: SR,SEF

NICU:

Adu, Ped: SR, SEF

Neo: N/A

General, OR, ICU,CCU: BIS Trend

NICU:

Adu: BIS Trend

Ped: BIS Trend

Neo: N/A

General, OR, ICU,CCU: All

NICU:

Adu: All

Ped: All

Neo: N/A

General, OR, ICU,CCU: BIS L

NICU:

Adu: BIS L

Ped: BIS L

Neo: N/A

General, OR, ICU,CCU: EMG

NICU:

Adu: EMG

Ped: EMG

Neo: N/A

70

70

Parameter 2 Yes Yes

BIS High Yes

BIS Low Yes

C.1.17

NMT

NMT from Mindray NMT module

Item Name

Stimulation mode

Configurable

In Config Mode

Yes

Interval Yes

Stimulation Current

Pulse Width

Stimulation Beep Volume

Block Recovery

DBS

NMT parameter switch

Yes

Yes

Yes

Yes

Yes

No

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Default

TOF

TOF, DBS

1 min

ST

0.1 Hz

Supra

200 μs

2

Off

DBS 3.3

On

Passport 12m/Passport 17m Operator's Manual C-15

Flow Scale

Vol Scale

Display Loop

Reference Loop

RR High

RR Low

PEEP High

PEEP Low

PIP High

PIP Low

MVe High

MVe Low

NMT from TOF-Watch® SX monitor

Item Name

Configurable

In Config Mode

Yes Alarm Sound

C.1.18

RM

Item Name

Alarm

Alm Lev

Apnea Delay

Configurable

In Config Mode

Yes

Yes

Yes

Sensor Type

TV/MV

Flow/Vol

Sweep

RR Source

No

Yes

Yes

Yes

No

Paw Scale Yes

Yes

Yes

No

No

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Default

Off

Default

0

40

1

30.0

2.0

On

Med

Adu, Ped:

Neo:

Disposable

TV

Flow

6.25 mm/s

Auto

Adu, Ped:

Neo:

Adu, Ped:

Neo:

Adu, Ped:

Neo:

PV Loop

On

Adu, Ped:

Neo:

Adu, Ped:

Neo:

10

20S

15 s

30

100

8

30

40

N/A

60

N/A

1200

N/A

C-16 Passport 12m/Passport 17m Operator's Manual

C.2

Routine Configuration

C.2.1

Alarm

Configurable

Item Name

In Config Mode In Monitor Mode

Alm Volume Yes Yes

Reminder Vol

Recording Length

Apnea Delay

Yes

Yes

Yes

Yes

Yes

Yes

Alarm Delay

ST Alarm Delay

VTac Latching

C.2.2

Screens

Yes

Yes

No

Yes

Yes

Yes

Item Name

Choose Screen

Display the ST segments on ECG screen

1

Select Wave Sequence for Normal Screen 6

7

8

4

5

2

3

Select Parameters for

Big Numerics Screen

9

10

Parameter 1

Parameter 2

Parameter 3

Parameter 4

Configurable

In Config Mode In Monitor Mode

Yes Yes

Yes

Yes

Yes

Yes

Yes

Yes

Default

General:

OR:

ICU, NICU, CCU:

Low

16 s

Adu, Ped:

Neo:

6 s

30 s

On

Default

Normal Screen

Unselected

ECG1

ECG2

SpO

2

+PR

Any IBP

Any IBP

CO

2

Paw

Flow/Vol

BIS

Resp

ECG

SpO

2

+PR

Resp

NIBP

20s

15 s

2

1

2

Passport 12m/Passport 17m Operator's Manual C-17

Item Name

C

O.M

M

General

OR

ICU

NICU

CCU

User Defaults

Item Name

C

O.M

M

General

OR

ICU

NICU

CCU

User Defaults

Select QuickKeys (Passport 12m)

Yes

No

NIBP Measure

Stop All

Zero IBP

Review

Standby

Screens

Patient Setup

Manual

Event

Realtime Print

Volume Setup

NIBP Measure

Stop All

→ oxyCRG

Review

Standby

Screens

Patient Setup

Manual

Event

Realtime Print

Volume Setup

NIBP Measure

Stop All

Zero IBP

Review

Standby

Screens

Patient Setup

Manual

Event

Realtime Print

Volume Setup

Select QuickKeys (Passport 17m)

Yes

No

NIBP Measure

Stop All

Zero IBP

Screens

Patient Setup

Manual Event

Realtime

Print

Print Setup

Minitrends

Volume Setup

Load Configuration

Privacy Mode

NIBP Measure

Stop All

Zero IBP

Screens

Patient Setup

Manual Event

Realtime

Print → Print Setup → Minitrends → Volume Setup → Load Configuration → PAWP

NIBP Measure

Stop All

Zero IBP

Screens

Patient Setup

Manual Event

Realtime

Print

Print Setup

Minitrends

Volume Setup

Load Configuration

Privacy Mode

NIBP Measure → Stop All → oxyCRG → Screens → Patient Setup → Manual Event → Realtime

Print

Minitrends

Zero IBP

Volume Setup

Load Configuration

Privacy Mode

NIBP Measure

Stop All

Zero IBP

Screens

Patient Setup

Manual Event

Realtime

Print

Print Setup

Minitrends

Volume Setup

Load Configuration

Privacy Mode

C-18 Passport 12m/Passport 17m Operator's Manual

C.2.3

Parameter/Wave Color

Item Name

Configurable

In Config Mode In Monitor Mode

ECG

NIBP

SpO

2

SpO

2 b

Δ

SpO

2

PR

TEMP

Art/Ao/UAP/FAP

/BAP/LV/P1~P4 (arterial pressure)

PA

CVP/ICP/P1~P4 (venous pressure)

LAP

RAP

UVP

CO

2

/tcpCO

2

RESP

AA

N

2

O

O

2

/tcpO

2

Hal

Enf

Iso

Des

Sev

C.O.

Paw

Flow/Vol

NMT

EEG L/BIS L Trend

EEG R/BIS R Trend

SvO

2

ScvO

2

CCO

No Yes

Default

Green

White

Cyan

Purple

Yellow

Cyan

White

Red

Blue

Green

Red

Orange

Purple

Cyan

Yellow

White

Yellow

Blue

Purple

Orange

Cyan

Yellow

Yellow

Yellow

Blue

White

Yellow

Blue

Cyan

Purple

Yellow

Passport 12m/Passport 17m Operator's Manual C-19

C.2.4

Review

Item Name

Tabular Trends

Interval

Full Disclosure

Trend Group

Trend Group

Graphic Trends Zoom

Waves

Save Waves

Gain

Sweep

C.2.5

Event

Item Name

Waveform 1

Waveform 2

Configurable

In Config Mode

No

No

No

No

No

No

No

No

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Default

General: 30 min

OR: 5 min

ICU, NICU, CCU: 30 min

Standard

Standard

90 min

2

Save ECG1 by default. x 1

25 mm/s

Configurable

In Config Mode

No

No

In Monitor Mode

Yes

Yes

Yes

Default

II

General, OR, ICU, CCU:

NICU

General, OR, ICU, CCU:

NICU

I

Pleth

Pleth

Resp

Waveform 3

C.2.6

Record

Item Name

Length

Interval

Paper Speed

IBP Overlap

Alm Rec X

X represents a parameter label.

No

No

No

No

Configurable

In Config Mode In Monitor Mode

No Yes

No Yes

Yes

Yes

Yes

Default

8 s

Off

25 mm/s

Off

Off

C-20 Passport 12m/Passport 17m Operator's Manual

C.2.7

Print

Item Name

Configurable

In Config Mode

No

No

No

No

No

No

No

No

No

No

Default

Paper Size

Print On Both Sides

ECG Reports

Tabular Trends

Reports

Amplitude

Sweep

Auto Interval

12-Lead Format

Set as End Case Report

Back

Resolution

Report Format

Currently Displayed Trended

Parameters

No

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes Selected

Graphic Trends

Reports

Realtime Reports

Standard Parameter Group No

Custom No

Not Print Blank Pages

Set as End Case Report

No

No

Back

Paginal Time

Set as End Case Report

No

No

No

Sweep

Select Wave

C.2.8

Night Mode

No

No

Item Name

Brightness

Alm Volume

QRS Volume

Key Volume

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Configurable

In Config Mode In Monitor Mode

Yes Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes Yes NIBP End Tone

C.2.9

Others

Item Name

Brightness

Key Volume

Configurable

In Config Mode In Monitor Mode

No Yes

No Yes

Default

1

0

1

2

Off

Default

5

2

Unselected

Unselected

Selected

Unselected

Auto

Auto

Unselected

Auto

Current

Letter

Off

10 mm/mV

25 mm/s

Off

12 x 1

Unselected

Auto

Auto

Parameter Oriented

Passport 12m/Passport 17m Operator's Manual C-21

C.3

User Maintenance Items

Item Name

Changing Pat. Loc.

Network Encrypt Switch

Barometric Pressure

Height Unit

Weight Unit

ST Unit

Press. Unit

Hb Unit

CVP Unit

CO

2

Unit

O

2

Unit

Temp Unit tcpCO

2

/ tcpO

2

Unit

Network Type

Address Type

Select CMS

ADT Query

Latching Alarms

Alarm Pause Time

No

No

No

No

No

No

No

No

No

No

No

No

No

No

Configurable

In Config Mode

No

No

No

No

No

Minimum Alarm Volume No

Alarm Sound

High Alarm Interval (ISO mode)

Medium (ISO mode)

Low Alarm Interval (ISO mode)

Reset Other Bed’s Alarms

Alarm Reset By Other Bed

Other Bed Disconnection Alarm

Reminder Interval

Reminder Tone

ECGLeadOff Lev.

SpO

2

SensorOff Lev.

IBPSensorOff Lev.

Lethal Arrh. OFF

Extended Arrh.

Alarm Light on Alarm Reset

Alarm Delay

ST Alarm Delay

Vtac Latching Lock

Wave Line

C-22

No

No

No

No

No

No

No

No

Yes

No

No

No

No

No

No

No

No

No

No

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Default

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

20

20

Off

On

On

3 min

On

Low

Low

Med

Disable

Enable

On

6 s

30 s

Enable

Med

Protected

Off

760 mmHg cm kg mV mmHg g/dl cmH

2

O mmHg

%

°C mmHg

LAN

Manual

Off

On

Lethal

2 min

General, ICU, NICU, CCU:

OR:

ISO

10

2

1

Passport 12m/Passport 17m Operator's Manual

Item Name

Primary Button

ECG Standard

Notch Freq.

Data Transfer Method

Transferred Data Length

Apply Module Settings

Parameter Switch

SpO

2

Tone

Clear CMS IP at startup

Signal Type

Nurse Call Setup

Contact Type

Alm Lev

Alarm Cat.

Configurable

In Config Mode

Yes

No

Yes

No

No

Yes

Yes

No

No

No

No

No

No

Yes

Yes

Yes

Yes

Yes

Yes

Yes

In Monitor Mode

Yes

Yes

Yes

Yes

Yes

Yes

Default

Left

AHA

50 Hz

Off

4 h

Off

Selected

Mode 1

Off

Continuous

Normally Closed

High, Med, Low

Phys., Tech.

Passport 12m/Passport 17m Operator's Manual C-23

FOR YOUR NOTES

C-24 Passport 12m/Passport 17m Operator's Manual

D

Alarm Messages

This chapter lists only the most important physiological and technical alarm messages. Some messages appearing on your monitor may not be included.

In this chapter:

The “I” column indicates how indications of technical alarms perform after the alarm system is reset: “A” means that some technical alarms are cleared; “B” indicates that some technical alarms are changed to the prompt messages; and “C” indicates that a “

” appears before the alarm message, appears in the alarm symbol area, and the indication of the alarm lamp depends on the alarm light setting. Refer to section

7.8 Resetting Alarms

for details.

The “Level” field indicates the alarm level: H means high, M means medium and L means low. “*” means the alarm level is user-adjustable.

XX represents a measurement or parameter label or module, such as ECG, NIBP, HR, ST-I, PVCs, RR, SpO

2

, PR, etc.

In the “Cause and Solution” column, corresponding solutions are given instructing you to troubleshoot problems. If the problem persists, contact your service personnel.

D.1

Physiological Alarm Messages

Measurement

XX

ECG

Alarm messages

XX Too High

XX Too Low

ECG Weak Signal

Asystole

VFib/VTac

Vtac

Vent. Brady

Extreme Tachy

Extreme Brady

R on T

Run PVCs

Couplet

PVCs/min

Bigeminy

Trigeminy

Tachy

Brady

M*

M*

M*

M*

M*

M*

M*

M*

H

H

H

H

H

H

Level Cause and solution

M* XX value has risen above the high alarm limit or fallen below the low alarm limit. Check the patient’s condition and check if the patient

M* category and alarm limit settings are correct.

H

The ECG signal is too weak for the monitor to perform ECG analysis.

Check the patient’s condition and the ECG connections.

Arrhythmia has occurred to the patient. Check the patient’s condition and the ECG connections.

Passport 12m/Passport 17m Operator's Manual D-1

CO

2

AG

RM

AG

Measurement

Resp

SpO

2 tcGas

NMT

Alarm messages

Missed Beats

Afib

Irr. Rhythm

Vent. Rhythm

Multif. PVC

Nonsus. Vtac

Pause

Pacer Not Pacing

Pacer Not Capture

Resp Apnea

Resp Artifact

SpO

2

Desat

SpO

2 b Desat

No Pulse

CO

2

Apnea

AG Apnea

RM Apnea

FiO

2

Too Low

+tcpCO

2

Alarm

+tcpO

2

Alarm

+SpO

2

Alarm

+PR Alarm

TOF Alarm

M*

M*

M*

M*

H

H

H

Level Cause and solution

M*

M*

M*

M*

M*

M*

M*

M*

M*

The pacer appears abnormal. Check the pacer.

H

H

H

H

The respiration signal was too weak for the monitor to perform respiration analysis. Check the patient’s condition and the Resp connections.

The patient’s heartbeat has interfered with his respiration. Check the patient’s condition and the Resp connections.

The SpO

2

or SpO

2 b value has fallen below the desaturation alarm limit. Check the patient’s condition and check if the alarm limit settings are correct.

The pulse signal was too weak for the monitor to perform pulse analysis. Check the patient’s condition, SpO

2 sensor and measurement site.

H

The patient stops breathing, or the respiration signal was too weak for the monitor to perform respiration analysis. Check the patient’s condition and the RM connections.

Check the patient’s condition, the ventilated O

2 content and the AG connections.

Parameter value has risen above the high alarm limit or fallen below the low alarm limit. Check the patient’s condition and check if the patient category and alarm limit settings are correct.

M*

For TOF

-

Watch SX monitor only. TOF value has risen above the high alarm limit or fallen below the low alarm limit. Check the patient’s condition and check if the patient category and alarm limit settings are correct.

D-2 Passport 12m/Passport 17m Operator's Manual

D.2

Technical Alarm Messages

Measurement

XX

MPM

T1/N1

ECG

Alarm message

XX SelfTest Err

XX Init Err

XX Init Err N(N is between 1 to

8)

XX Comm Err

XX Comm Stop

XX Limit Err

XX Overrange

MPM 12V Err

MPM 5V Err

Transfer Module battery to be protected and not work.

ECG Noisy

ECG High Freq. Noise

H

H

H

Level

H

H

L

L

H

H

H

I

C

A

A

A

C

C

C

C

C

C

Transfer Module battery aged. Replace the battery

L C

Transfer Module No Battery H C

PWR interrupted. Check meas. state

High Technical Alarm

Medium Technical Alarm

L

H

M

A

C

C

Low Technical Alarm

ECG Lead Off

L

L*

C

B

ECG YY Lead Off L* B

Note: YY represents the leadwires, V (V1, V2, V3, V4, V5,

V6,), LL, LA, RA, as per AHA standard, or C (C1, C2, C3,

C4, C5, C6), F, L and R as per IEC standard.

L

L

A

A

Cause and solution

An error occurred to the XX module, or there is a problem with the communications between the module and the monitor.

Re-plug the module and restart the monitor, or plug the module into another monitor.

XX parameter limit is accidentally changed.

Contact your service personnel.

The measured XX value is not within the specified range for XX measurement. Contact your service personnel.

An error occurred to the power supply part of the MPM module. Contact your service personnel.

The battery will be soon protected and will not supply power. If you are going to use

T1/N1 for patient transport, please replace the battery.

Replace the battery.

T1/N1 has no battery. Install a battery for

T1/N1.

Power supply failed accidently. Check the measurements when the monitor restarts.

T1/N1 has a high/medium/low technical alarm. Check the T1/N1 monitor for the alarm.

The electrode has become detached from the patient or the lead wire has become disconnected from the adapter cable. Check the connections of the electrodes and leadwires.

The ECG signal is noisy. Check for any possible sources of signal noise around the cable and electrode, and check the patient for great motion.

High frequency signals are detected on the

ECG analysis lead. Check for any possible source of interference around the cable and electrode.

Passport 12m/Passport 17m Operator's Manual D-3

Measurement

Resp

Temp

SpO

D-4

2

Alarm message

ECG Low Freq. Noise

ECG Amplitude Too Small L

ECG Config. Err L

Resp Disturbed

Temp Cal. Err

T1 Sensor Off

T2 Sensor Off

SpO

2

Sensor Off

SpO

2 b Sensor Off

SpO

2

Sensor Fault

SpO

2 b Sensor Fault

SpO

2

No Sensor

SpO

2 b No Sensor

SpO

2

Unknown Sensor

SpO

2 b Unknown Sensor

SpO

2

Sensor Incompatible

SpO

2 b Sensor Incompatible

SpO

2

Too Much Light

SpO

2 b Too Much Light

L

L

L

L

L

H

L

L

L*

L

SpO

2

Low Signal

SpO

2

b Low Signal

SpO

2

No Pulse

SpO

2 b No Pulse

SpO

2

Interference

SpO

2 b Interference

SpO

2

Comm Abnormal

L

L

L

H

Level

L

A

C

A

A

I

A

C

C

Cause and solution

Low frequency signals are detected on the

ECG analysis lead. Check for any possible source of interference around the cable and electrode.

The ECG amplitude didn’t reach the detected threshold. Check for any possible source of interference around the cable and electrode.

ECG configuration is wrongly downloaded.

Check the downloaded configuration and re-download the correct configuration.

The respiration circuit is disturbed. Restart the monitor.

A calibration failed. Restart the monitor.

The Temp sensor has become detached from the patient or the module. Check the sensor connections.

B

C

B

C

The SpO

2

sensor has become detached from the patient or the module, or there is a fault with the SpO

2

sensor, or an unspecified SpO

2 sensor has been used. Check the sensor application site and the sensor type, and make sure if the sensor is damaged.

Reconnect the sensor or use a new sensor.

C

C

C

C

C

A

There is too much light on the SpO

2

sensor.

Move the sensor to a place with lower level of ambient light or cover the sensor to minimize the ambient light.

The SpO

2

signal is too low or too weak. Check the patient’s condition and change the sensor application site. If the error persists, replace the sensor.

The SpO

2

sensor failed to obtain pulse signal.

Check the patient’s condition and change the sensor application site. If the error persists, replace the sensor.

The SpO

2

signal has been interfered. Check for any possible sources of signal noise around the sensor and check the patient for great motion.

An error occurred to the SpO

2

measurement module, or there is a problem with the

Passport 12m/Passport 17m Operator's Manual

Measurement

NIBP

Alarm message Level I Cause and solution

SpO

SpO

SpO

SpO

2

2

2

2 b Comm Abnormal

Board Fault b Board Fault b has been closed

L

H

C

C communications between the module and the monitor. Re-plug the module and restart the monitor, or plug the module into another monitor.

There is a problem with the SpO

2 measurement board. Do not use the module and contact your service personnel.

Different types of SpO

2

measurement modules are applied. Use the same type of

SpO

2

measurement modules.

The NIBP cuff is not properly connected, or there is a leak in the airway.

NIBP Loose Cuff

NIBP Air Leak

NIBP Pneumatic Leak

L

L

L

A

A

A

NIBP Cuff Type Wrong

NIBP Air Pressure Err

NIBP Weak Signal

NIBP Signal Saturated

NIBP Overrange

L

L

L

L

L

A

A

A

A

A

Check the NIBP cuff and pump for leakages.

The cuff type applied mismatches the patient category. Verify the patient category and replace the cuff.

An error occurred to the air pressure. Verify that the monitor application site meets the environmental requirements and check if there is any source that affects the air pressure.

The patient’s pulse is weak or the cuff is loose.

Check the patient’s condition and change the cuff application site. If the error persists, replace the cuff.

The NIBP signal is saturated due to excess motion or other sources.

The measured NIBP value is not within the specified range.

NIBP-XX Over Upper Limit L A The measured pressure is greater than the specified NIBP measurement upper limit.

NIBP-XX Over Lower Limit L A The measured pressure is lower than the specified NIBP measurement lower limit.

XX represents diastolic pressure, mean pressure, or systolic pressure.

NIBP Excessive Motion

NIBP Cuff Overpress.

NIBP Equip Err

NIBP Timeout

NIBP Measure Failed

L

L

H

L

L

A

A

A

A

A

Check the patient’s condition and reduce the patient motion.

The NIBP airway may be occluded. Check the airway and measure again.

An error occurred during NIBP measurement and therefore the monitor cannot perform analysis correctly. Check the patient’s condition and NIBP connections, or replace the cuff.

Passport 12m/Passport 17m Operator's Manual D-5

Measurement

IBP

C.O.

ScvO

D-6

2

Alarm message

NIBP Illegally Reset

YY Sensor Off

YY Disconnected H

YY Sensor Fault

YY No Pulse

YY represents an IBP label.

TB Sensor Off

Optical Module Err

ScvO

2

Signal Too High

ScvO

2

Signal Too Low

ScvO

2

Too Much Light

L

Optical Module Disconnected L

ScvO

ScvO

2

2

Comm Abnormal

ScvO

2

Comm Err

ScvO

2

Init Err

Unsupported CeVOX version

Comm Stop

Disconnect from Vigilance

H

H

H

L

L

L

L

Level

L

M*

M

L

H

H

L

C

A

I

A

A

C

C

C

C

A

A

C

A

Cause and solution

An illegal reset occurred during NIBP measurement. Check if the airway is occluded.

Check the sensor connection and reconnect the sensor.

The liquid way is disconnected from the patient, or the three-way valve is open to the air. Check the connection of the liquid way, or check the valve is open to the patient. If the problem remains, contact the Customer

Services Dept. for help.

Replace the sensor.

The catheter may be occluded. Please flush the catheter.

Check the sensor connection and reconnect the sensor.

Check the module connection. Change a module if necessary.

Check the sensor and reposition the catheter, then recalibrate the sensor.

Check and reposition the catheter, then recalibrate the sensor. Avoid the backlight which is excessively strong.

Connect the optical module.

Remove/connect the module again or restart the machine. If the problem remains, contact the Customer Services Dept. for help.

A

A

A

A

C

Remove/connect the module again. If the problem remains, contact the Customer

Services Dept. for help.

The module version is not compatible with the system. Please contact the Customer

Services Dept. for help.

Remove/connect the module again or restart the machine. If the problem remains, contact the Customer Services Dept. for help.

The Vigilance II monitor is disconnected from the host monitor

Check the connection between the Vigilance

II monitor and the host monitor.

Passport 12m/Passport 17m Operator's Manual

Measurement Alarm message

Disconnects from Vigileo

Disconnect From EV1000 L

CO

2

Sensor High Temp

CO

2

Sensor Low Temp

CO

2

Temp Overrange

CO

2

Airway High Press.

CO

2

Airway Low Press.

CO

2

High Barometric Press.

L

L

L

L

L

L

CO

2

CO

2

Low Barometric Press.

CO

2

FilterLine Occluded

CO

2

No Watertrap

CO

2

Check Adapter

CO

2

FilterLine Err

CO

2

Zero Failed

CO

2

System Err

CO

2

Check Cal.

Passport 12m/Passport 17m Operator's Manual

L

L

L

L

L

L

L

L

Level

L

C

C

C

A

A

C

C

B

A

C

I

C

C

C

C

C

C

Cause and solution

The Vigileo TM monitor is disconnected from the host monitor.

Check the connection between the Vigileo

TM monitor and the host monitor.

The EV1000 monitor is disconnected from the host monitor.

In this case, check the connection between the EV1000 monitor and the host monitor.

When EV1000 monitor works in VolumeView mode, finishes measurement, and then admits a new patient, this alarm is issued.

In this case, perform one measurement after admitting a new patient at the EV1000 monitor.

When the information of a new patient is not entered into the EV1000 within 30 seconds, this alarm is issued.

In this case, enter the patient information within 30 seconds.

Check, stop using or replace the sensor.

Check, stop using or replace the sensor.

The operating temperature of the CO

2 module goes beyond the specified range.

After it restores within the specified range, the module will restart automatically.

An error occurred in the airway pressure.

Check the patient connection and patient circuit, and then restart the monitor.

Check the CO

2

connections, make sure that the monitor application site meets the requirements, and check for special sources that affect the ambient pressure. Restart the monitor.

The airway or watertrap was occluded. Check the airway and remove the occlusion.

Check the watertrap connections.

There is a problem with the airway adapter.

Check, clean or replace the adapter.

Check if there is a leak in the CO

2

sample line or the CO

2

sample line has been occluded.

Check the CO

2

connections. After the sensor’s temperature becomes stabilized, perform a zero calibration again.

Re-plug the module or restart the monitor.

Perform a calibration.

D-7

Measurement Alarm message

CO

2

Check Airway

CO

2

No Filterline

CO

2

No Sensor

CO

2

Main Board Err

CO

2

Checking Sensor

CO

2

Replace Scrubber&Pump

CO

2

15V Overrange

CO

2

Hardware Err tcGas Low Battery

Level

L

L

H

L

H

L

L

H

M tcGas tcGas Battery Depleted H

TCM Temperature Too High H

TCM Alert L

AG

AG No Watertrap

AG Change Watertrap

AG Watertrap Type Wrong

O

2

Accuracy Unspecified

N

2

O Accuracy Unspecified

CO

2

Accuracy Unspecified

Enf Accuracy Unspecified

Iso Accuracy Unspecified

Sev Accuracy Unspecified

Hal Accuracy Unspecified

Des Accuracy Unspecified awRR Accuracy Unspecified

AG Hardware Err

AG Airway Occluded

AG Zero Failed

NMT (Mindray)

NMT No Main Cable

NMT Sensor Fault

L

L

L

L

L

L

L

L

L

L

L

L

L

H

L

L

L

I

C

C

C

A

C

C

A

C

C

A

C

A

A

A

A

A

A

A

A

A

A

A

A

B

A

A

Cause and solution

An error occurred to the airway.

Make sure that the filterline is connected.

Make sure that the sensor is connected.

There is a problem with the CO

The measured value has exceeded the specified accuracy range.

Remove the AG module. Stop using the module and contact your service personnel.

Check the airway and remove the occlusion.

Re-plug the module or restart the monitor, and then perform a zero calibration again.

Check that NMT patient cable is properly connected to the NMT module.

The NMT sensor has a fault. Reconnect or replace the sensor.

2

module.

Re-plug the module or restart the monitor.

C

C

C

Connect the TCM monitor or SenTec Digital

Monitor with AC mains.

TCM monitor or SenTec Digital Monitor has less than 5 minutes running time on battery.

Connect the TCM monitor or SenTec Digital

Monitor with AC mains immediately.

The temperature in TCM CPU is too high.

Please shut down TCM monitor immediately.

A TCM technical alarm is presented. Please check the TCM monitor to identify the cause of alarm.

Check the connections of the watertrap and re-connect it.

Wait until the change is completed.

Make sure that a correct watertrap has been used.

D-8 Passport 12m/Passport 17m Operator's Manual

Measurement Alarm message

NMT No Sensor

NMT Stimulation Electrode Off L

NMT Stimulation Current Over

Limit

NMT Power Err

L

H

NMT Abnormal Reset

NMT

(TOF-Watch

®

SX)

TWSX Low Battery

TWSX Battery Depleted

TWSX No Acceleration Sensor

TWSX No Temp Sensor

TWSX No Stimulation Cable

TWSX Bad Electrode

Connection

L

L

M

H

L

L

RM

TWSX Technical Alarm

RM No Sensor

RM Sensor Reversed

RM Zero Failed

RM Power Err

BIS High Imped.

BIS Sensor Off

BIS DSC Err

BIS

BIS DSC Malf

BIS No Cable

BISx Disconnected

BIS No Sensor

BIS Wrong Sensor Type

BIS Sensor Too Many Uses

SQI<50%

SQI<15%

BIS Sensor Expired

BIS Sensor Fault

Passport 12m/Passport 17m Operator's Manual

L

L

L

L

L

L

L

L

L

L

L

L

L

L

L

L

L

L

Level

L

L

I

A

C

A

C

C

A

A

A

C

C

A

A

A

A

A

A

A

A

C

A

C

A

A

B

B

C

C

B

B

Cause and solution

Check that NMT sensor is properly connected to the NMT patient cable. If the alarm persists, replace the sensor.

Check that NMT sensor is properly connected to the NMT patient cable. If the alarm persists, check the application of electrodes.

The output stimulation current exceeds the specification. Contact your service personnel.

Contact your service personnel.

Take out the NMT module and plug it again in the module rack. Restart the monitor or test the module with another Passport m Series monitor. If the problem persists, contact your service personnel.

Replace the battery.

Replace the battery.

Connect the acceleration sensor.

Connect the temperature sensor.

Connect the stimulation cable.

Reattach the electrode.

An NMT technical alarm is presented. Please check the TOF-Watch® SX monitor to identify the cause of alarm.

Check and reconnect the sensor.

Perform a zero calibration again.

There is a problem with the power supply.

Re-plug the module or restart the monitor.

Check and reconnect the BIS sensor.

An error occurred to the DSC during receiving signals. Check the DSC.

The DSC automatically shuts down as a result of malfunction. Check the DSC.

Check the BIS cables.

Check the BISx module.

Check the BIS sensor.

Check or replace the sensor.

Replace the sensor.

The SQI value is too low. Check the patient’s condition and the sensor connections.

Replace the sensor.

Re-attach or Replace BIS Sensor

D-9

Measurement Alarm message

Disconnect/Reconnect BIS

12V Too High

12V Too Low

5V Too High

5V Too Low

3.3V Too High

3.3V Too Low

Battery Too Low

Level

H

H

H

H

H

L

H

H

Power

Recorder

System

D-10

Battery Overload

Transfer Module battery to be protected and not work.

Transfer Module battery aged. Replace the battery iView operates obnormally.

Please use AC.

RT Clock Need Reset

RT Clock Not Exist

Recorder Init Err N

N is within 1 to 8.

Recorder SelfTest Err

Recorder Comm Err

Recorder S. Comm Err

Recorder Unavailable

Recorder Vlt High

Recorder Vlt Low

Recorder Head Hot

Rec Paper Wrong Pos.

IP Address Conflict

No CMS

H

H

L

H

L

L

L

L

H

L

L

L

L

L

L

L

L

I

C

C

C

C

C

C

C

C

Cause and solution

Re-plug the BIS Module

There is a problem with the system power supply. Restart the monitor.

Connect the monitor to an AC power source and allow the batteries to charge.

C

Too many parameter modules are connected, causing system overload and high power consumption. Use AC power supply.

C

A

C

C

C

C

T1 /N1 battery will be soon protected and will not supply power. If you are going to use the

T1 /N1 for patient transport, please replace the battery.

T1 /N1 battery lifetime is expired.

Replace the battery with a new one.

When batteries are used as the power source, iView system can not properly work. If you want to use iView system, please power the monitor with an AC power source.

Internal backup battery cell fails. Contact your service personnel.

Contact your service personnel.

Restart the monitor.

A

A

A

A

C

C

C

A

A

A

Stop the recording and restart the monitor.

An error occurred to the system power supply. Restart the monitor.

The recorder has been working for too long time. Stop the recording and resume the recording till the recorder’s printhead cools down.

Re-load the recorder paper.

Set a new IP address.

The monitor is disconnected from the CMS.

Check network connection.

Passport 12m/Passport 17m Operator's Manual

Measurement Alarm message

Other Bed Disconnected

PWR interrupted. Check meas. state

Restoring Last Config. Failed

Level

L

L

L

Loading Default Config. Failed.

USB Drive Err

Storage Card Err

Storage Card Space Low

USB Drive Space Low

L

M

M

L

L

I

A

A

A

C

B

A

A

A

Cause and solution

Check network connection.

Power supply failed accidently. Check the measurements when the monitor restarts.

Restart the monitor. If the problem persists, there may be an EEPROM failure. Contact your service personnel.

Restart the monitor. If the problem persists, there may be an EEPROM failure. Contact your service personnel.

Disconnect the USB memory and reconnect it properly.

If the problem persists, format the USB memory.

If the problem still persists, replace the USB drive.

Restart the monitor. If the problem persists, format the SD card.

The storage card has abnormal data. Format the storage card.

Delete unnecessary data from the USB memory, or replace the USB memory.

Passport 12m/Passport 17m Operator's Manual D-11

FOR YOUR NOTES

D-12 Passport 12m/Passport 17m Operator's Manual

E

SpO

2

Sensor Accuracy

A.

1

Accuracy of Masimo SpO

2

Sensors

Table information for the plots below show ARMS values measured with Masimo SET Oximetry Technology in a clinical study.

DCI/DCIP/4050/4051

Passport 12m/Passport 17m Operator's Manual E-1

E-2

4053

Inf/Neo/NeoPt/4002/4003/4004/4005

Adtx/Pdtx/4000/4001

Passport 12m/Passport 17m Operator's Manual

A.2

Accuracy of

Nellcor

SpO

2

Sensors

 SpO

2

Accuracy for Nellcor TM Sensors vs. Co-Oximeters

SpO

2

Range 100% to 70% 100% to 90%

DS-100A 1.64% 1.16%

OXI-P/I, OXI-A/N

MAX-AI, MAX-PI, MAX-II

MAX-NI

2.41%

1.62%

1.85%

1.38%

1.49%

1.71%

Modified Bland-Altman for SpO

2

-DS-100A Sensor: (SpO

2

- SaO

2

) vs. SaO

2

90% to 80%

1.67%

2.50%

1.57%

1.51%

80% to 70%

2.25%

3.60%

2.50%

1.59%

Modified Bland-Altman for SpO

2

-OXI-P/I, OXI-A/N Sensors: (SpO

2

- SaO

2

) vs. SaO

2

Passport 12m/Passport 17m Operator's Manual E-3

Modified Bland-Altman for SpO

2

- MAX-AI, MAX-AI, MAX-PI, MAX-II Sensors: SaO

2

vs. (SpO

2

- SaO

2

)

Modified Bland-Altman for SpO

2

- MAX-NI Sensor: SaO

2

vs. (SpO

2

- SaO

2

)

E-4 Passport 12m/Passport 17m Operator's Manual

F

Symbols and Abbreviations

F.1

Symbols

mcg mEq mg min ml

L lb m mAh

Mb mm mmHg cmH2O ms mV mW

MΩ h

Hz in kg kPa

DS

ºF g

GHz

GTT bps

ºC cc cm dB

µA

µV

µs

A

Ah bpm microampere microvolt

Microsecond ampere ampere hour beat per minute bit per second centigrade cubic centimeter centimeter decibel dyne second fahrenheit gram gigahertz gutta hour hertz inch kilogram kilopascal litre pound meter milliampere hour mega byte microgram milli-equivalents milligram minute milliliter millimeter millimeters of mercury centimeters of water millisecond millivolt milliwatt megaohm

Passport 12m/Passport 17m Operator's Manual F-1

±

×

<

>

+

=

%

/

V

VA

Ω

W nm rpm s nanometer breath per minute second volt volt ampere ohm watt minus, negative percent per; divide; or; none; not applicable plus equal to less than greater than less than or equal to greater than or equal to plus or minus multiply

F.2

Abbreviations

F-2

ATMP aVF aVL aVR awRR

BAP

Base Flow

BC

BIS

AaDO2

AAMI

AC

ACI

Adu

AG

AHA air cyl.

Air Flow

ANSI

Ao

Art alveolar-arterial oxygen gradient

Association for Advancement of Medical Instrumentation alternating current acceleration index adult anaesthesia gas

American Heart Association

Air cylinder pressure air flow

American National Standard Institute aortic pressure arterial

Barometric pressure left foot augmented lead left arm augmented lead right arm augmented lead airway respiratory rate brachial arterial pressure base flow burst count bispectral index

Passport 12m/Passport 17m Operator's Manual

EDV

EE

EEC

EEG

EMC

EMG

EMI

Enf

ESU

Dia

DPI dPmx

DVI

DO

2

DO

2

I

ECG

CP

CPI

CPO

Cstat

CVP

DC

Des

CISPR

CMOS

CMS

C.O.

CO

2

COHb

Compl

CCO

CaO

2

CCO

CCU

CE

CFI

CIS

BP

BPSK

BSA

BT

BTPS

C.I.

CCI

Cdyn

Passport 12m/Passport 17m Operator's Manual blood pressure binary phase shift keying body surface area blood temperature body temperature and pressure, saturated cardiac index

Continuous Cardiac Index dynamic compliance

Continuous Cardiac Output arterial oxygen content continuous cardiac output cardiac (coronary) care unit

Conformité Européenne cardiac function index

Clinical Information System

International Special Committee on Radio Interference complementary metal oxide semiconductor central monitoring system cardiac output carbon dioxide carboxyhemoglobin compliance cardiopulmonary cardiac power index

Cardiac Power Output static compliance central venous pressure direct current desflurane diastolic dot per inch left ventricular contractility digital video interface oxygen delivery oxygen delivery index electrocardiograph end-diastolic volume

Energy expenditure

European Economic Community electroencephalogram electromagnetic compatibility electromyography electromagnetic interference enflurane electrosurgical unit

F-3

F-4

ESV

EtIso

EtSev

EtCO

2

EtN

2

O

EtO

EtO

2

ELWI

ESVI

Et

EtAA

EtAA 2nd

EtDes

EtEnf

EtHal

FiN

2

O

FiO

2

Flow fmand

FPGA

FRC

FiAA 2nd

FiDes

FiEnf

FiHal

FiIso

FiSev

FiCO

2

FreqMIN fsigh fSIMV

EVLW

Exp%

Exp. Flow

Exp. MAC f

FAP fapnea

FCC fCMV

FDA

FEV1.0%

FG

Fi

FiAA end systolic volume end systolic volume index end-tidal

End-tidal anesthetic agent

2nd Exp. Agent end-tidal anesthetic agent end-tidal carbon dioxide end-tidal nitrous oxide ethylene oxide end-tidal oxygen extravascular lung water index extravascular lung water inspiration termination level expiratory flow

Expired minimum alveolar concentration breath rate femoral arterial pressure breath rate for apnea ventilation

Federal Communication Commission

CMV frequency

Food and Drug Administration first second forced expiratory volume ratio

Fresh gas flow fraction of inspired

Inspired anesthetic agent

2nd Insp. Agent inspired anesthetic agent fraction of inspired carbon dioxide fraction of inspired nitrous oxide fraction of inspired oxygen flow mandatory breathing frequency field programmable gate array

Fractional residual capacity minimum breath frequency sigh rate frequency of SIMV

Passport 12m/Passport 17m Operator's Manual

Iso

IT

ITBI

ITBV

LA

LAP

Lat

LCD

LCW

LCWI

Leak Comp

LED

LL

LVD

LVDS

LVET

IBP

IBW

ICP

ICT/B

ICU

ID

I:E

IEC

IEEE

Ins

Insp.Flow

Insp. MAC

△ int.PEEP

IP

Hct

Hb

Hb-CO

HbO

2

HIS

HR

I:E fspn ftot

F-Trigger

FV

GEDV

GEDI

GEF

Hal

Passport 12m/Passport 17m Operator's Manual spontaneous breathing frequency total breath rate inspiratory trigger level (flow trigger) flow-volume global end diastolic volume global end diastolic volume index global ejection fraction halothane haematocrit hemoglobin carbon mono-oxide hemoglobin oxyhemoglobin hospital information system heart rate inspiratory-expiratory ratio invasive brood pressure ideal body weight intracranial pressure intracranial catheter tip pressure transducer intensive care unit identification inspiratory time: Expiratory time ratio

International Electrotechnical Commission

Institute of Electrical and Electronic Engineers inspired minimum inspiration flow

Inspired minimum alveolar concentration intermittent PEEP internet protocol isoflurane injectate temperature

Intrathoracic Blood Volume Index

Intrathoracic Blood Volume left arm left atrial pressure lateral liquid crystal display left cardiac work left cardiac work index leak compensation light emitting diode left leg low voltage directive low voltage differential signal left ventricular ejection time

F-5

F-6

MRN

MV

MVe

MVi

MVLEAK

MVspn

N/A

N

2

N

2

O

N

2

O cyl.

N

2

O Flow

Neo

NIBP

NIF

LVSW

LVSWI

MAC

Art mean

MDD

MetHb

%MinVol

MRI

NMT

O

2

Δ

O

2

O

2

%

O

2

CI

O

2

cyl.

O

2

cyl.2nd

O

2

EI

O

2

Flow

O

2

R

OR oxyCRG

PA

Pair

Papnea pArt-D pArt-M pArt-S

Paux Mean

Paux Min

Paux Peak

Paw

PAWP left ventricular stroke work left ventricular stroke work index minimum alveolar concentration mean arterial pressure

Medical Device Directive methemoglobin

Percentage of minute volume to be delivered magnetic resonance imaging medical record number minute volume expiratory minute volume inspiratory minute volume leakage minute volume spontaneous breathed minute volume not applied nitrogen nitrous oxide

N

2

O cylinder pressure

N

2

O flow neonate noninvasive blood pressure negative inspiratory force neuromuscular transmission oxygen

Difference between inspiratory and expiratory O2 oxygen concentration oxygen consumption index

Oxygen cylinder pressure

Secondary oxygen cylinder pressure oxygen extraction index

O

2

flow oxygen extraction ratio operating room oxygen cardio-respirogram pulmonary artery

Air supply pressure apnea pressure diastolic artery pressure mean artery pressure systolic artery pressure

Mean auxiliary pressure

Minimum auxiliary pressure

Peak auxiliary pressure airway pressure pulmonary artery wedge pressure

Passport 12m/Passport 17m Operator's Manual

P0.1

P0.1 time

R

RA

RAM

Ramp

RAP

RAW

RCexp

Ppeak

Pplat

PPV

PR

Psupp

PTC

PTP

P-Trigger

PVC

PVR

PVRI

PVPI pArt pCVP

PD

Pleth

Plimit

Plow

Pmax

Pmean

PN

2

O

PO

2

PEEPtot

PEF

PEP

Phigh

PIF

Pinsp

PIP

Peak Flow

Ped

PEEP

PEEP/CPAP

PEEPe

PEEPi

PEEPi time

Passport 12m/Passport 17m Operator's Manual photodetector peak flow pediatric positive end expiratory pressure

PEEP/CPAP

Extrinsic positive end-expiratory pressure intrinsic positive end-expiratory pressure

Intrinsic PEEP age (elapsed time since last maneuver) total PEEP peak expiratory flow pre-ejection period upper pressure level peak inspiratory flow pressure control level of inspiration peak inspiratory pressure plethysmogram pressure limit level lower pressure level maximum airway rressure mean pressure

N

2

O supply pressure

Oxygen supply pressure peak pressure plateau pressure

Pulse Pressure Variation pulse rate pressure support level post tetanic count

Pressure time product inspiratory trigger level (pressure trigger) premature ventricular contraction pulmonary vascular resistance pulmonary vascular resistance index pulmonary vascular permeability index artery pressure central venous pressure

100 ms occlusion pressure

P0.1 age (elapsed time since last maneuver) right right arm random access memory

Ramp right atrial pressure airway resistance

Expiratory time constant

F-7

F-8

TB

TD

Temp

Texp

TFC

TFI

SVV

SvO

2

ScvO

2

Sync

Sys

Tapnea

Taxil

TFT

Thigh

Ti max

SQI

SR

STR

SV

SVI

SVR

SVRI

SaO

2

SEF

Sev

SFM

SI

SMR

SpO

2

RCinsp

RL

RM

RQ

RR

RSBI

Rstat

RVEF

Rdyn

Re

Rec

Resp

RHb

Ri

Rise Time%

Inspiratory time constant dynamic lung resistance expiratory resistance record, recording respiration reduced hemoglobin inspiratory resistance rise time right leg respiratory mechanics

Respiratory quotient respiration rate rapid shallow breathing index static lung resistance right ventricular ejection fraction arterial oxygen saturation spectral edge frequency sevoflurane self-maintenance stroke index satellite module rack arterial oxygen saturation from pulse oximetry signal quality index suppression ratio systolic time ratio stroke volume

Stroke Volume Index systemic vascular resistance systemic vascular resistance index stroke volume variation mixed venous oxygen saturation central venous oxygen saturation synchronization systolic pressure apnea interval axillary temperature

Blood Temperature temperature difference temperature

Expiratory time thoracic fluid content thoracic fluid index thin-film technology time for the upper pressure level maximum inspiration time

Passport 12m/Passport 17m Operator's Manual

VO

2

/kg

VO

2

/m 2

VO

2

I

VTe/TVe

VTi/TVi

VT

VTapnea

VTe spn

VTsigh

WLAN

WOB

WOBimp

UVP

VAC

VCO

2

VEPT

VI

VO

2

Trise

Tslope

Tube ID

TVe i

UAP

UPS

Tinsp

Tip

TIP:TI

Tlow

TOF

Toral

TP

Tplat

TRC

Trect

Trigger

Trig

Window time of inspiration

Inspiratory pause time percentage of inspiratory plateau time in inspiratory time time for the lower pressure level train of four oral temperature total power plateau time

Tube resistance compensation rectal temperature trigger sensitivity trigger window rise time time for the pressure to rise to target pressure tube ID expiratory tidal volume inspiratory tidal volume umbilical arterial pressure uninterruptible power supply umbilical venous pressure volts alternating current

CO

2

production volume of electrically participating tissue velocity index oxygen consumption

Oxygen consumption per body weight

Oxygen consumption per body surface area oxygen consumption index expiratory tidal volume inspiratory tidal volume tidal volume apnea tidal volume spontaneous expiratory tidal volume sigh tidal volume wireless local area network work of breathing imposed work of breathing

Passport 12m/Passport 17m Operator's Manual F-9

FOR YOUR NOTES

F-10 Passport 12m/Passport 17m Operator's Manual

P/N: 046-005013-00 (1 7 .0)

advertisement

Related manuals

Download PDF

advertisement

Table of contents