Traceable 4366 TDS & Conductivity Meter Operating instructions
Below you will find brief information for 4366. This meter measures conductivity and total dissolved solids (TDS). It features automatic temperature compensation, selectable normalization, and a user-adjustable TDS factor. The device offers manual range selection and a hold function for easy recording of measurements.
Advertisement
Advertisement
SPECIFICATIONS
Conductivity—
Range: 0 to 1999 µS (micromhos)
Resolution: 1
Range: 0.00 to 19.99 mS (mhos)
Resolution: 0.01
Total Dissolved Solids—
Range: 0 to 1999 ppm (parts per million)
Resolution: 1
Range: 0.00 to 19.99 ppt (parts per thousand)
Resolution: 0.1
TDS Factor: 0.40 to 1.00, 0.01 increments
Automatic Temperature Compensation—
Range: 32.0 to 122.0 °F / 0.0 to 50.0 °C
Normalization: 20 °C or 25 °C (user selectable)
Coefficient: 0.0 to 4.0% per °C, 0.1 increments
Temperature Measurement—
Range: 32.0 to 122.0 °F / 0.0 to 50.0 °C
Resolution: 0.1°
OPERATION
This section contains basic operating instructions, to insure accurate results and for more advanced functions, refer to the additional sections that follow.
1. Remove the protective cover from the probe head.
3. Press the
Ç
MODE
measure (conductivity or TDS).
ON/OFF
SET
button.
button to select the desired unit of
◄
4. Place the probe head into the solution to be measured.
5. To insure Automatic Temperature Compensation,
immerse the probe head and temperature sensor
completely. For accurate Automatic Temperature
Compensation, the temperature sensor must reach
equilibrium with the liquid being measured, this could
take several minutes.
6. Insure that there is adequate movement of the liquid
around the tip of the probe head (0.2 - 0.3 meters per
second minimum). This can be achieved by stirring the
meter in the liquid.
The meter will automatically select the range that is the most appropriate for the value being measured. For example, when measuring a 500 µS solution, the meter will select the µS range and display 500 µS as opposed to displaying 0.50 mS. To manually select the range, press and hold the Ç
MODE
button for approximately three
(3) seconds, if the value being measured is able to be displayed in the alternate range, the meter will switch ranges (see the “Range Mode Selection” section).
7. Take the desired measurements.
8. Press the ON/OFF
SET
button to turn the meter off.
To preserve battery life, always turn the meter off when
not in use.
9. Rinse the probe head with distilled/deionized water and
replace the protective cover. Always replace the
protective cover to prevent damage to the probe head.
If no button is pressed for twenty (20) minutes, the meter will turn off automatically to preserve battery life (see the
“Automatic Shutoff” section).
POWER ON DISPLAY SEQUENCE
When the meter is turned on (by pressing the ON/OFF
SET button), the display will cycle through showing the following items:
All Segments - Display check tnr Normalization temperature value (see the
“Selecting Temperature Normalization Value”
section) tco Temperature Compensation Coefficient value
(see the “Setting Temperature Compensation
Coefficient” section).
tds TDS Factor value (see the “Setting TDS
Factor” section).
ran Range display mode (see the “Range Mode
Selection” section).
CONDUCTIVITY CALIBRATION
One calibration point may be entered for each of the measurement ranges (0 to 1999 µS and 0.00 to 19.99 mS).
To insure accuracy:
–Always use Traceable
–Select a Traceable ®
® Conductivity Standards
Conductivity Standard with a value as close as possible to the value of the unknown sample.
–If possible, the Traceable ® Conductivity Standard and unknown sample should be at the same temperature.
The ideal temperature is 25 °C.
Previous calibration points are cleared when the meter is calibrated to a new value.
1. Remove the protective cover from the probe head.
2. Turn the meter on by pressing the ON/OFF
SET
3. Place the probe head into the Traceable ®
button.
Conductivity
Standard.
4. To insure Automatic Temperature Compensation,
immerse the probe head and temperature sensor
completely. For accurate Automatic Temperature
Compensation, the temperature sensor must reach
equilibrium with the liquid being measured, this could
take several minutes.
5. Insure that there is adequate movement of the liquid
around the tip of the probe head (0.2 - 0.3 meters per
second minimum). This can be achieved by stirring the
meter in the liquid.
6. Press and hold the
three (3) seconds, CAL will appear at the top of the
display and the conductivity measurement will flash.
7. Press the
Ç
MODE
È
HLD/CAL
and/or
button for approximately
È
HLD/CAL
buttons to adjust the
display to the know value of the Traceable ® Conductivity
Standard. The meter may only be adjusted by ±30%
of the detected value.
8. With the the know value of the Traceable
Standard appearing and CAL no longer flashing,
press the ON/OFF
SET
button, the meter will return to the
measurement display.
® Conductivity
VIEWING CALIBRATION DATA
1. Turn the meter on by pressing the
2. Press and hold the ON/OFF
SET
ON/OFF
SET
button.
button for approximately
three (3) seconds, P1.0
will appear at the bottom of
the display.
3. Press the
4. Press the
Ç
MODE
Ç
button three (3) times, P4.0
will
appear at the bottom of the display.
ON/OFF
SET
button, P4.1
will appear at the
bottom of the display. The top of the display will show
the value that the meter was calibrated to for the
range of 0 to 1999 µS. (If no user calibration has been
performed, “---” will appear.)
5. Press the
MODE
button, P4.2
will appear at the
bottom of the display. The top of the display will
show the value that the meter was calibrated to for the
range of 0 to 19.99 mS. (If no user calibration has
been performed, “---” will appear.)
6. Press the ON/OFF
SET
button, P4.0 will appear at the
bottom of the display.
7. Press and hold the ON/OFF
SET
measurement display.
button for approximately
three (3) seconds, the meter will return to the
SELECTING °F OR °C
1. Turn the meter on by pressing the
2. Press and hold the ON/OFF
SET
ON/OFF
SET
button.
button for approximately
three (3) seconds, P1.0
will appear at the bottom of
the display.
3. Press the
4. Press the
switch to the opposite unit of measure. If the meter
had been displaying °F, it will now display °C, and
vice-versa. Each press of the
ON/OFF
SET
ON/OFF
SET
Ç
MODE
button, the temperature unit of
button, the temperature reading will
Ç
MODE
button will switch
the display to the opposite unit of measure.
5. With the desired unit of measure appearing, press the
button three (3) times, P1.0 will appear at the
bottom of the display.
6. Press and hold the ON/OFF
SET
measurement display.
button for approximately
three (3) seconds, the meter will return to the
The unit of measure selected will be the default unit of measure until changed.
RANGE MODE SELECTION
The meter will automatically select the range that is the most appropriate for the value being measured. For example, when measuring a 500 µS solution, the meter will select the µS range and display 500 µS as opposed to displaying 0.50 mS.
To manually select the display range, press and hold the Ç
MODE
button for approximately three (3) seconds, if the value being measured is able to be displayed in the alternate range, the meter will switch ranges.
Each time the Ç
MODE
button for pressed and held, the range mode will cycle through the available modes:
ATO Automatic range selection
1
2
0 to 1999 µS / 0 to 1999 ppm
0.00 to 19.99 mS / 0.00 to 19.99 ppt
HOLD FUNCTION
The hold function allows the readings on the display to be
“frozen” so that they may be recorded.
1. Press and release the
È
È
appear at the top of the display.
2. Once the reading has been recorded, press and
release the
HLD/CAL
button to return the display to
the current reading. “HOLD” will no longer
appear on the display.
HLD/CAL
button to “freeze”
the display at the current reading. “HOLD” will
AUTOMATIC SHUTOFF (Disable)
If no button has been pressed for twenty (20) minutes, the Automatic Shutoff feature will turn the meter off to preserve battery life.
To disable the Automatic Shutoff feature:
1. With the meter turned off, press and hold the
button, then press and hold the
“ n ” appears on the display.
2. Release the
be disabled.
È
HLD/CAL
and ON/OFF
SET
ON/OFF
SET
È
HLD/CAL
button until
buttons. The meter
will power on and the Automatic Shutoff feature will
The Automatic Shutoff feature is automatically enabled when the meter is powered off and turned back on.
SELECTING TEMPERATURE NORMALIZATION
VALUE
Conductivity is greatly influenced by temperature. Most fluids increase in conductivity as temperature increases.
Depending on the value selected here, Automatic
Temperature Compensation (ATC) will automatically make corrections to the reading and display a value as if the sample was 25 °C or 20 °C, no matter what the actual sample temperature is. The international standard temperature normalization value is 25 °C.
The following procedure is used to change the temperature normalization value from 25 °C to 20 °C, or from 20 °C to 25 °C.
1. Turn the meter on by pressing the
2. Press and hold the
Ç
button.
button for approximately
three (3) seconds, P1.0
will appear at the bottom of
the display.
3. Press the ON/OFF
SET
ON/OFF
SET
ON/OFF
SET
button two (2) times, the temperature
normalization value will flash on the display.
4. Press the Ç
MODE
button, the temperature normalization
value will change. Each press of the
MODE
button will
switch between 25.0 °C (77.0 °F) and 20.0°C (68.0°F).
5. With the desired temperature normalization value
appearing, press the
P1.0 will appear at the bottom of the display.
6. Press and hold the ON/OFF
SET
measurement display.
ON/OFF
SET
button two (2) times,
button for approximately
three (3) seconds, the meter will return to the
The temperature normalization value selected will be the default value until changed.
SETTING TEMPERATURE COMPENSATION
COEFFICIENT
Conductivity is greatly influenced by temperature. Most fluids increase in conductivity as temperature increases.
The value set here will determine how much the
Automatic Temperature Compensation (ATC) feature will adjust the value for every 1 °C in order to automatically make corrections to the reading and display a value as if the sample was 25 °C or 20 °C (see the “Selecting
Temperature Normalization Value” section), no matter what the actual sample temperature is.
Although every fluid will have a different temperature coefficient, a temperature compensation coefficient of
2.0% is a generally accepted standard compensation value. If the temperature coefficient of the solution being measured is known, that value should be entered in order to obtain the most accurate results. To take nontemperature compensated measurements, a temperature compensation coefficient of 0.0% should be entered.
The following procedure is used to set the temperature compensation coefficient.
1. Turn the meter on by pressing the
2. Press and hold the
È
button.
button for approximately
three (3) seconds, P1.0
will appear at the bottom of
the display.
3. Press the ON/OFF
SET
ON/OFF
SET
ON/OFF
SET
button three (3) times, the
temperature coefficient value will flash on the display.
4. Press the Ç
MODE
and/or
HLD/CAL
buttons to adjust the
display to the desired temperature coefficient value.
5. With the desired temperature coefficient value
appearing, press the
P1.0 will appear at the bottom of the display.
6. Press and hold the ON/OFF
SET
measurement display.
ON/OFF
SET
button one (1) time,
button for approximately
three (3) seconds, the meter will return to the
The temperature coefficient value selected will be the default value until changed.
SETTING TDS FACTOR
The Total Dissolved Solids (TDS) value that appears on the display is a result of the conductivity measurement of the fluid being multiplied by the TDS factor.
Every liquid will have a different TDS factor. An approximation of the dissolved solids for a fresh-water source can be arrived at by using a TDS factor of 0.66. If the TDS factor of the solution being measured is known, that value should be entered in order to obtain the most accurate results.
The following procedure is used to set the TDS factor.
1. Turn the meter on by pressing the
2. Press and hold the
MODE
ON/OFF
SET
three (3) seconds, P1.0
will appear at the bottom of
the display.
3. Press the
bottom of the display.
4. Press the
5. Press the
Ç
MODE
ON/OFF
SET
on the display.
Ç and/or È
HLD/CAL
ON/OFF
SET
display to the desired TDS factor value.
button.
button for approximately
button, P2.0
will appear at the
button, the TDS factor value will flash
buttons to adjust the
6. With the desired TDS factor value appearing, press the
ON/OFF
SET
display.
button, P2.0 will appear at the bottom of the
7. Press and hold the ON/OFF
SET
measurement display.
button for approximately
three (3) seconds, the meter will return to the
The TDS factor value selected will be the default value until changed.
RESET TO FACTORY DEFAULTS
Resetting the meter will clear all calibration data and reset the user selectable values to the factory defaults.
Factory defaults are as follows:
Temperature display: °C
Normalization Temperature: 25 °C
Temperature Coefficient: 2.1%
TDS Factor: 0.50
The following procedure is used to reset the meter to the factory default values.
1. Turn the meter on by pressing the
2. Press and hold the ON/OFF
SET
ON/OFF
SET
button.
button for approximately
three (3) seconds, P1.0
will appear at the bottom of
the display.
3. Press the
4. Press the
5. Press the
Ç
MODE
Ç
MODE
button two (2) times, P3.0
will
appear at the bottom of the display.
ON/OFF
SET
button, the n will flash on the display.
button, the y will flash on the display.
6. With y appearing, press the button, P3.0 will appear at the bottom of the display and the values
will have been reset to the factory defaults.
7. Press and hold the ON/OFF
SET
measurement display.
ON/OFF
SET
button for approximately
three (3) seconds, the meter will return to the
DISPLAY MESSAGES
Upper Display Messages:
--- Indicates that the meter is in the manual range selection mode and the value being measured is higher than 1999 µS or 1999 ppm.
E03 Indicates that the value of the solution being measured exceeds the measurement range of the unit.
E04 Indicates that the value of the solution being measured exceeds the measurement range of the unit based on a temperature measurement error (see the “Lower Display Messages” section).
Lower Display Messages:
E0
1
Indicates that the temperature sensor is damaged.
E0
2
Indicates that the temperature of the solution being measured is below the measurement range of the unit.
E0
3
Indicates that the temperature of the solution being measured is above the measurement range of the unit.
ALL OPERATION DIFFICULTIES
If this meter does not function properly for any reason, replace the batteries with a new, high quality batteries
(see the “Battery Replacement” section). Low battery power can occasionally cause an number of “apparent” operational difficulties. Replacing the batteries with new fresh batteries will solve most difficulties.
BATTERY REPLACEMENT
An erratic display, faint display, no display, or a battery symbol appearing on the display are all indicators that the batteries need replacement. The battery cover is located at the top of the unit. Unscrew the battery cover by turning it counter-clockwise. Remove the exhausted batteries and replace them with four (4) new #357/LR44 size silver oxide batteries. Make certain to insert the new batteries with the proper polarity as indicated by the
“+” and “–” symbols in the battery compartment. The
“+” side of the silver oxide battery is the flat side with the engraving. Replace the battery cover and tighten securely.
WARRANTY, SERVICE, OR RECALIBRATION
For warranty, service, or recalibration, contact:
CONTROL COMPANY
12554 Old Galveston Rd. Suite B230
Webster, Texas 77598 USA
Ph. 281 482-1714 • Fax 281 482-9448
E-mail [email protected] www.traceable.com
Control Company is ISO 9001:2008 Quality-Certified by DNV and ISO/IEC 17025:2005 accredited as a
Calibration Laboratory by A2LA.
TRACEABLE
CONDUCTIVITY/
®
TOTAL DISSOLVED
SOLIDS
METER
INSTRUCTIONS
Control Company Cat. No. 4366
Traceable ® is a registered trademark of Control Company.
© 2016 Control Company. 92-4366-00 Rev. 1 061716
Download
Advertisement
Key features
Conductivity and TDS measurement
Automatic Temperature Compensation
User-selectable temperature normalization
Adjustable TDS factor
Manual range selection
Data hold function
Frequently asked questions
Place the probe in a Traceable® Conductivity Standard, press and hold the HLD/CAL button, and adjust the display to the known value using the MODE and HLD/CAL buttons.
With the meter off, press and hold the HLD/CAL button, then press and hold the ON/OFF button until 'n' appears on the display.
Enter the setup menu (hold ON/OFF), then press ON/OFF again, and use the MODE button to toggle between temperature units.