Urisys 1100

®

Urine Analyzer – Setup Instruction Sheet

For complete details regarding installation, please see Section 4 of Operator’s Manual.

1) Unpacking

Carefully remove the contents of the Urisys 1100 ® Urine Analyzer box and check for the following items: a) Urisys 1100 ® Urine Analyzer b) AC adapter c) Power cord d) Roll of printer paper e) Test strip tray, Type C, for reading Chemstrip ® 10 MD* Test Strips

a

d

e

b

c Items included, but not pictured.

• Operator’s Manual (including Policies and Procedures CD)

• User Training CD

• Quick Reference Guide

• Warranty Card

Please note: Chemstrip urine test strips and Chemstrip

Calibration Strips are not included and must be purchased separately.

2) Setup

Place analyzer on level surface.

Connect AC adapter to power cord and then to power socket at rear of analyzer. Plug power cord into power outlet.

3) Insert Strip Tray

Select appropriate test strip tray. (Use Type C to read

Chemstrip 10 MD* Test Strips. ) Hold strip tray with gray reference pad facing up. Slide test strip tray into slot below function keys until retaining bar closes.

4) Install Paper

Release printer cover by pressing area immediately below and to right of printer paper slot. Lift cover and place printer paper roll into compartment as shown below. Pull paper forward over the edge of paper slot. Close cover.

5) Self Check

Switch on Urisys 1100 ® Urine

Analyzer using on/off switch at rear of analyzer. The analyzer automatically performs a self check. If self check is completed successfully, the message “SELF

CHECK OK” prints along with time and date.

6) Calibrate

Before using the analyzer, it must be calibrated using Chemstrip Calibration

Strips (Cat. No. 11379194160). After the self check, the message “REPEAT

CALIBRATION” appears on the display.

Press the blue START button and the message “START CALIBRATION” displays.

Place a calibration strip, with the pads facing upward, on test strip tray. (The leading edge of strip should be held by the clip at end of test strip tray. The retaining bar must be open.) Press START button.

The display will then read, “Heat up! Please wait, xx sec left.” Once the countdown finishes, the measurement begins. If calibration results are acceptable, the message “CALIBRATION O.K.” displays and results are printed.

7) Begin Testing

The Urisys 1100 ® Analyzer is ready to begin reading test strips when the message “INSERT STRIP” is displayed.

Customer Support

If you need assistance, contact

Roche Diagnostics Customer

Support Center at 1-800-428-4674,

7 days a week, 24 hours a day, and

365 days per year.

* Hospitals may use Chemstrip ® 10 UA Urine Test Strips.

COBAS, LIFE NEEDS ANSWERS, URISYS 1100 and CHEMSTRIP are trademarks of Roche.

© 2012-2018 Roche Diagnostics. All rights reserved.

Rx Only

VV-06557-02 https://usdiagnostics.roche.com

03755550001(07) 2018-09 USA