Invacare Perfecto2 IRC5PAW, Perfecto2 IRC5PO2AW User Manual


Add to my manuals
32 Pages

advertisement

Invacare Perfecto2 IRC5PAW, Perfecto2 IRC5PO2AW User Manual | Manualzz

User Manual

Perfecto

2

Series

Oxygen Concentrators without SensO

2

®,

Oxygen Concentrators with SensO

2

HomeFill® Compatible

Model IRC5PO2AW

Model IRC5PO2VAW

Model IRC5PAW

Dealer: This manual MUST be given to the end user.

User: BEFORE using this product, read this manual and save for future reference.

For more information regarding

Invacare products, parts, and services, please visit www.invacare.com

WARNING

WARNING

DO   NOT   use   this   product   or   any   available   optional   equipment   without   first   completely   reading   and   understanding   these   instructions   and   any   additional   instructional   material   such   as   owner’s   manuals,   service   manuals   or   instruction   sheets   supplied   with   this   product   or   optional   equipment.

  If   you   are   unable   to   understand   the   warnings,   cautions   or   instructions,   contact   a   healthcare   professional,   dealer   or   technical   personnel   before   attempting   to   use   this   equipment  ‐  otherwise,   injury   or   damage   may   occur.

ACCESSORIES WARNING

Invacare   products   are   specifically   designed   and   manufactured   for   use   in   conjunction   with   Invacare   accessories.

  Accessories   designed   by   other   manufacturers   have   not   been   tested   by   Invacare   and   are   not   recommended   for   use   with   Invacare   products.

ACCESSORIES

There   are   many   different   types   of   humidifiers,   oxygen   tubing,   cannulas   and   masks   that   can   be   used   with   this   device.

  You   should   contact   your   local   home   care   provider   for   recommendations   on   which   of   these   devices   will   be   best   for   you.

  They   should   also   give   you   advice   on   the   proper   usage,   maintenance,   and   cleaning.

NOTE:   Updated   versions   of   this   manual   can   be   found   at   www.invacare.com.

Perfecto

2

™Series 2 Part No 1160877

TABLE OF CONTENTS

ACCESSORIES ......................................................2

SPECIAL NOTES ...................................................4

LABEL LOCATION ................................................6

SECTION 1—GENERAL GUIDELINES .....................7

Radio Frequency Interference...................................................................8

SECTION 2—FEATURES ......................................10

SECTION 3—HANDLING ....................................11

Unpacking.......................................................................................................11

Inspection.......................................................................................................11

Storage............................................................................................................11

SECTION 4—TECHNICAL DESCRIPTION .............12

SECTION 5—TYPICAL PRODUCT PARAMETERS .13

SECTION 6—OPERATING INSTRUCTIONS ..........17

Introduction...................................................................................................17

Select a Location..........................................................................................17

Set Up..............................................................................................................18

Flowrate..........................................................................................................21

SensO

2

Oxygen Purity Indicator - Model IRC5PO2AW................22

Initial Startup of the Concentrator ........................................................22

Explanation of Indicator Lights - Model IRC5PAW..........................23

Elapsed Time Meter (Hour Meter)........................................................23

SECTION 7—MAINTENANCE .............................24

Routine Maintenance..................................................................................24

SECTION 8—TROUBLESHOOTING GUIDE ............27

SECTION 9—OPTIONAL ACCESSORIES ...............29

RECYCLING INFORMATION ..............................30

LIMITED WARRANTY ........................................31

Part No 1160877 3 Perfecto2™Series

SPECIAL NOTES

SPECIAL NOTES

Signal   words   are   used   in   this   manual   and   apply   to   hazards   or   unsafe   practices   which   could   result   in   personal   injury   or   property   damage.

  Refer   to   the   following   table   for   definitions   of   the   signal   words.

SIGNAL WORD

DANGER

WARNING

CAUTION

MEANING

Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.

Warning indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.

Caution indicates a potentially hazardous situation which, if not avoided, may result in property damage or minor injury or both.

NOTICE

The   information   contained   in   this   document   is   subject   to   change   without   notice.

DANGER

DO NOT SMOKE while using this device. Keep all matches, lit cigarettes or other sources of ignition out of the room in which this product is located and away from where oxygen is being delivered.

NO SMOKING signs should be prominently displayed.

Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air. Failure to observe this warning can result in severe fire, property damage and cause physical injury or death.

Perfecto

2

™Series 4 Part No 1160877

SPECIAL NOTES

CAUTION

“Caution:   Statutory   law   can   restrict   this   device   to   sale   by   or   on   order   of   a   physician,   or   any   other   practitioner   licensed   by   the   law   of   the   governmental   agency   in   which   he/she   practices   to   use   or   order   the   use   of   this   device.”

Invacare   recommends   an   alternate   source   of   supplemental   oxygen   in   the   event   of   a   power   outage,   alarm   condition   or   mechanical   failure.

  Consult   your   physician   or   equipment   provider   for   the   type   of   reserve   system   required.

This   equipment   is   to   be   used   as   an   oxygen   supplement   and   is   not   considered   life   supporting   or   life   sustaining.

Part No 1160877 5 Perfecto

2

™Series

LABEL LOCATION

LABEL LOCATION

Perfecto

2

FOR OPERATING INSTRUCTIONS or ALARMS (See Operator’s Manual)

DANGER - RISK OF FIRE

NO SMOKING - Keep ALL sources of ignition out of the room in which this product is located and away from areas where oxygen is being delivered. Textiles, oil and other combustiles are easily ignited and burn with great intensity in oxygen enriched air.

DANGER - ELECTRIC SHOCK

DO NOT remove cover. Refer servicing to qualified service personnel.

Flowrates below 0.5 L/min will trigger Low Flow Alarm (rapid audible alarm).

Set Flowmeter at 2.5 L/min or less when filling HomeFill® cylinders.

HomeFill®

Compatible

Serial Number Label is located on the resonator intake assembly

Specification Label is located on the back near the bottom

Perfecto

2

™Series 6 Part No 1160877

SECTION 1—GENERAL GUIDELINES

SECTION 1—GENERAL GUIDELINES

In   order   to   ensure   the   safe   installation,   assembly   and   operation   of   the   Perfecto

2

  concentrator   these   instructions   MUST   be   followed.

WARNING

This   section   contains   important   information   for   the   safe   operation   and   use   of   this   product.

DANGER

Risk   of   electric   shock.

  DO   NOT   disassemble.

  Refer   servicing   to   qualified   service   personnel.

  No   user   serviceable   parts.

TO   REDUCE   THE   RISK   OF   BURNS,   ELECTROCUTION,  

FIRE   OR   INJURY   TO   PERSONS.

Avoid   using   while   bathing.

  If   continuous   usage   is   required   by   the   physician’s   prescription,   the   concentrator   MUST   be   located   in   another   room   at   least   2.5

  m   (7   ft)   from   the   bath.

DO   NOT   come   in   contact   with   the   concentrator   while   wet.

DO   NOT   place   or   store   product   where   it   can   drop   into   water   or   other   liquid.

DO   NOT   reach   for   product   that   has   fallen   into   water.

  Unplug  

IMMEDIATELY.

If   the   concentrator   has   a   damaged   cord   or   plug,   if   it   is   not   working   properly,   if   it   has   been   dropped   or   damaged,   or   dropped   into   water,   call   qualified   technician   for   examination   and   repair.

 

A   spontaneous   and   violent   ignition   may   occur   if   oil,   grease   or   greasy   substances   come   in   contact   with   oxygen   under   pressure.

 

These   substances   MUST   be   kept   away   from   the   oxygen   concentrator,   tubing   and   connections,   and   all   other   oxygen   equipment.

  DO   NOT   use   any   lubricants   unless   recommended   by   Invacare.

Avoid   creation   of   any   spark   near   medical   oxygen   equipment.

 

This   includes   sparks   from   static   electricity   created   by   any   type   of   friction.

Part No 1160877 7 Perfecto

2

™Series

SECTION 1—GENERAL GUIDELINES

Operating Information

For   optimum   performance,   Invacare   recommends   that   each   concentrator   be   on   and   running   for   a   minimum   of   30   minutes   at   a   time.

  Shorter   periods   of   operation   may   reduce   maximum   product   life.

 

Keep   the   oxygen   tubing,   cord,   and   unit   out   from   under   such   items   as   blankets,   bed   coverings,   chair   cushions,   clothing   and   away   from   heated   or   hot   surfaces,   including   space   heaters,   stoves   and   similar   electrical   appliances.

 

DO   NOT   move   or   relocate   concentrator   by   pulling   on   the   power   cord.

NEVER   drop   or   insert   any   object   or   liquid   into   any   opening.

Invacare   recommends   that   Crush ‐ Proof   oxygen   tubing   be   used   with   this   product   and   not   exceed   15.2

  m   (50   ft)   in   length.

There   are   no   user   serviceable   parts.

  This   does   not   include   normal   maintenance   items.

  See   maintenance   section   for   user   maintenance   items.

A   product   should   NEVER   be   left   unattended   when   plugged   in.

 

Make   sure   the   Perfecto

2

  is   Off   when   not   in   use.

Close   supervision   is   necessary   when   this   product   is   used   near   children   or   physically ‐ challenged   individuals.

Additional   monitoring   or   attention   may   be   required   for   patients   using   this   device   who   are   unable   to   hear   or   see   alarms   or   communicate   discomfort.

 

DO   NOT   connect   the   concentrator   in   parallel   or   series   with   other   oxygen   concentrators   or   oxygen   therapy   devices.

Radio Frequency Interference

This   equipment   has   been   tested   and   found   to   comply   with   EMC   limits   specified   by   IEC/EN   60601 ‐ 1 ‐ 2.

  These   limits   are   designed   to   provide   a   reasonable   protection   against   electromagnetic   interference   in   a   typical   medical   installation.

Perfecto

2

™Series 8 Part No 1160877

SECTION 1—GENERAL GUIDELINES

Other   devices   may   experience   interference   from   even   the   low   levels   of   electromagnetic   emissions   permitted   by   the   above   standards.

  To   determine   if   the   emissions   from   the   Perfecto

2

  is   causing   the   interference,   turn   the   Perfecto

2

  Off.

  If   the   interference   with   the   other   device(s)   stops,   then   the   Perfecto

2

  is   causing   the   interference.

  In   such   rare   cases,   interference   may   be   reduced   or   corrected   by   one   of   the   following   measures:

• Reposition,   relocate,   or   increase   the   separation   between   the   equipment.

• Connect   the   equipment   into   an   outlet   on   a   circuit   different   from   that   to   which   the   other   device(s)   is   connected.

Part No 1160877 9 Perfecto

2

™Series

SECTION 2—FEATURES

SECTION 2—FEATURES

Oxygen Outlet

Oxygen Purity Indicator

Lights / Fault and Power

Indicator Lights

Flowmeter

Circuit Breaker

Elapsed Time

Meter

Power Switch

REAR VIEW

Power

Cord

Perfecto

2

™Series

Cabinet

Filter

*HF Outlet

Fitting

*NOTE:   This   outlet   fitting   is   to   be   used   only   for   filling   oxygen   cylinders   with   the  

HomeFill   home   oxygen   compressor.

  The   outlet   fitting   does   not   affect   concentrator   performance.

 

Refer   to   the   HomeFill   owner ʹ s   manual,   part   number   1116444,   for   connection   and   operating   instructions.

  When   not   in   use,   the   plug   provided   with   the   concentrator   should   be   inserted   into   the   outlet   fitting.

  For   more   information   about   the  

HomeFill,   contact   your  

Invacare   dealer.

10 Part No 1160877

SECTION 3—HANDLING

SECTION 3—HANDLING

The   concentrator   should   ALWAYS   be   kept   in   the   upright   position   to   prevent   cabinet   damage   while   being   transported.

If   the   concentrator   is   to   be   reshipped   by   common   carrier,   additional   cartons   are   available   from   Invacare.

Unpacking

NOTE:   For   this   procedure,   refer   to   FIGURE   3.1.

1. Check   for   any   obvious   damage   to   the   carton   or   its   contents.

  If   damage   is   evident,   notify   the   carrier,   or   your   local   dealer.

2. Remove   all   loose   packing   from   the   carton.

3. Carefully   remove   all   the   components   from   the   carton.

FIGURE 3.1 Unpacking

NOTE:   Unless   the   oxygen   concentrator   is   to   be   used   immediately,   leave   concentrator   in   its   packaging   for   storage   until   use   of   the   concentrator   is   required.

Inspection

1. Inspect/examine   exterior   of   the   oxygen   concentrator   for   nicks,   dents,   scratches   or   other   damages.

  Inspect   all   components.

Storage

1. Store   the   repackaged   oxygen   concentrator   in   a   dry   area.

2. DO   NOT   place   objects   on   top   of   repackaged   concentrator.

Part No 1160877 11 Perfecto

2

™Series

SECTION 4—TECHNICAL DESCRIPTION

SECTION 4—TECHNICAL

DESCRIPTION

The   Invacare   Perfecto

2

  concentrator   is   used   by   patients   with   respiratory   disorders   who   require   supplemental   oxygen.

  The   device   is   not   intended   to   sustain   or   support   life.

 

The   oxygen   concentration   level   of   the   output   gas   ranges   from  

87%   to   95.6%.

  The   oxygen   is   delivered   to   the   patient   through   the   use   of   a   nasal   cannula.

 

The   Invacare   Perfecto

2

  concentrator   uses   a   molecular   sieve   and   pressure   swing   adsorption   methodology   to   produce   the   oxygen   gas   output.

  Ambient   air   enters   the   device,   is   filtered   and   then   compressed.

  This   compressed   air   is   then   directed   toward   one   of   two   nitrogen   adsorbing   sieve   beds.

 

Concentrated   oxygen   exits   the   opposite   end   of   the   active   sieve   bed   and   is   directed   into   an   oxygen   reservoir   where   it   is   delivered   to   the   patient.

The   Invacare   Perfecto

2

  concentrator   is   capable   of   operation   by   the   patient   in   a   home   environment   or   in   an   institutional   environment.

  Device   operates   at   a   nominal   230   VAC/50   Hertz   supply.

Service   information   will   be   available   upon   request   to   qualified   technical   personnel   only.

Perfecto

2

™Series 12 Part No 1160877

SECTION 5—TYPICAL PRODUCT PARAMETERS

SECTION 5—TYPICAL PRODUCT

PARAMETERS

Alternating Current

Type BF equipment

Unit running

Unit not running

Attention - Consider Accompanying Documents

DO NOT smoke

Class II, Double Insulated

IPX1

Electrical

Requirements:

Protected against dripping water

230 VAC ± 10% (253 VAC/207 VAC), 50 Hz

Rated Current Input: 1.4 A

Sound Level:

Altitude:

39.5 dBA Average

All Perfecto

2

Models: Up to 8,000 ft (2438 meters) above sea level without degradation of concentration levels. From 8,000 ft to 13,129 (4000 meters) below 90% efficiency.

All Perfecto

2

Models while filling a HomeFill System:

Up to 6,000 ft (1828 meters) above sea level without degradation of concentration levels. From 6,000 ft (1828 meters) to

13,129 ft (4000 meters) below 90% efficiency

Oxygen Output

Concentration Levels:

NOTE:   Stated   concentration   levels   achieved   after   initial   warm ‐ up   period  

(approximately   30   minutes).

87% to 95.6% at 0.5 to 5 L/min

93% ± 3% at 2 L/min

Part No 1160877 13 Perfecto

2

™Series

SECTION 5—TYPICAL PRODUCT PARAMETERS

Maximum Outlet

Pressure:

Flow Range:

Low Flow Alarm:

34.5 kPa ± 3.45 kPa (5 psi ± 0.5 psi)

0.5 to 5 L/min (maximum)

For flowrates less than 1 L/min, we recommend the use of the

Invacare Pediatric Flowmeter Accessory (IRCPF16AW)

0 L/min to 0.5 L/min

Rapid Audible Alarm Beeping (No Accessories Connected)

Average Power

Consumption: Perfecto

2

Perfecto

2

VAW - 320 W@ 5 L/min

Pressure Relief

Mechanism

Operational at:

Change in maximum recommended flow when back pressure of 7kPa is applied:

Filters:

Safety System:

Width:

Height:

Depth:

Weight:

Shipping Weight:

241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)

0.7 L/min

Cabinet, Outlet HEPA and Compressor Inlet

Current overload or line surge shutdown

High temperature compressor shutdown

High Pressure Alarm w/compressor shutdown

Low Pressure Alarm w/compressor shutdown

Battery Free Power Loss Alarm

SensO

2

Oxygen System (SensO

2

Model) Low Flow Alarm

38.1 cm ± 1 cm (15 in ± 3/8 in)

58.4 cm ± 1 cm (23 in ± 3/8 in)

30.5 cm ± 1 cm (12 in ± 3/8 in)

Perfecto

2

AW Models - 20.5 kg ± 1 kg (45 lbs ± 2 lbs)

Perfecto

2

VAW Models - 18.1 kg ± 1 kg (40 lbs ± 2 lbs)

Perfecto

2

Perfecto

2

AW Models - 22.7 kg ± 1 kg (50 lbs ± 2 lbs)

VAW Models - 20.4 kg ± 1 kg (45 lbs ± 2 lbs)

Operating Ambient

Temperature:

Exhaust Cooling Air

Temperature:

Oxygen Output

Temperature:

10°C - 35°C (50°F - 95°F) at 20-60% relative humidity

Less than Ambient +19°C (+ 35°F)

Less than Ambient +3°C(+ 6°F)

Perfecto

2

™Series 14 Part No 1160877

SECTION 5—TYPICAL PRODUCT PARAMETERS

Cabinet:

No AP/APG

Impact Resistant flame-retardant plastic cabinet that conforms to UL

94-V0

Not suitable for use in the presence of a flammable anaesthetic mixture

Standards and

Regulatory Listing: IRC5PO2AW/IRC5PAW/IRC5PO2VAW:

IEC/EN60601-1, A1, A2

IEC/EN60601-1-2

IEC/EN61000-3-2

IEC/EN61000-3-3,

CE marked model

Electrical:

Placement:

IRC5PO2AW/IRC5PO2VAW Models Only

ISO8359

MDD 93/42/EEC, Annex I and IX

IRC5PO2AW & IRC5PO2VAW

No extension cords

No closer than 7.5 cm (3 in) from any wall, furniture, draperies, or furniture to assure sufficient air flow

Avoid deep pile carpets and heaters, radiators or hot air registers

Floor location only

No confined spaces (Example: No closets)

Tubing: 2 m (7 ft) cannula with a maximum 15 m (50 ft) of Crush-Proof

Tubing (DO NOT pinch)

Time of Operation: Up to 24 hours per day

Recommended

Storage and Shipping

Temperature:

Environment:

-29°C to 65°C (-20°F to 150°F) at 15-95% relative humidity

Smoke and soot-free

No confined spaces (Example: No closets)

Minimum Operating

Time: 30 Minutes

Part No 1160877 15 Perfecto

2

™Series

SECTION 5—TYPICAL PRODUCT PARAMETERS

O

2

IRC5PO2AW & IRC5PO2VAW INDICATORS

LABEL

SYMBOL

O

2

PURITY INDICATOR LIGHTS

(LED)

GREEN Indicator Light SYSTEM OKAY

O

2

over 85% (± 2%)

O

2

Between 73% (± 3%) to 85% (±2%)

SYSTEM FAILURE

O

2

Below 73% (±3%)

YELLOW Indicator light

A. YELLOW Solid

B. YELLOW Flashing Sensor

Failure

Call a qualified technician.

RED Indicator Light

Continuous Audible Alarm

Sieve-GARD Compressor

Shutdown

LABEL

SYMBOL

I/O

IRC5PAW INDICATORS

STATUS

SYSTEM OKAY

INDICATOR LIGHTS

(LED)

GREEN Indicator Light

SYSTEM FAILURE RED Indicator light

Continuous Audible Alarm

Sieve-GARD™Compressor

Shutdown

Call a qualified technician.

Perfecto

2

™Series 16 Part No 1160877

SECTION 6—OPERATING INSTRUCTIONS

SECTION 6—

OPERATING

INSTRUCTIONS

Introduction

Your   oxygen   concentrator   is   intended   for   individual   use.

  It   is   an   electronically   operated   device   that   separates   oxygen   from   room   air.

  It   provides   high   concentration   of   oxygen   directly   to   you   through   a   nasal   cannula.

  Clinical   studies   have   documented   that   oxygen   concentrators   are   therapeutically   equivalent   to   other   types   of   oxygen   delivery   systems.

Your   provider   will   show   you   how   to   use   your   oxygen   concentrator.

  He/She   should   be   contacted   with   any   questions   or   problems   regarding   your   oxygen   concentrator.

  This   owner’s   manual   will   tell   you   about   your   concentrator   and   will   serve   as   a   reference   as   you   use   your   concentrator.

Select a Location

WARNING

NEVER   block   the   air   openings   of   the   product   or   place   it   on   a   soft   surface,   such   as   a   bed   or   couch,   where   the   air   opening   may   be   blocked.

  Keep   the   openings   free   from   lint,   hair   and   the   like.

Keep   unit   at   least   7.5

  cm   (3   in)   away   from   walls,   draperies,   furniture,   and   the   like.

You   may   select   a   room   in   your   house   where   using   your   oxygen   concentrator   would   be   most   convenient.

  Your   concentrator   can   be   easily   rolled   from   room   to   room   on   its   casters.

Part No 1160877 17 Perfecto

2

™Series

SECTION 6—OPERATING INSTRUCTIONS

Your   oxygen   concentrator   will   perform   best   when   operated   under   the   conditions   outlined   in   the   Typical   Product  

Parameters   on   page 13.

  Usage   in   environments   other   than   those   described   may   result   in   the   need   for   increased   equipment   maintenance.

  The   air   intake   of   the   unit   should   be   located   in   a   well   ventilated   area   to   avoid   airborne   pollutants   and/or   fumes.

  DO   NOT   use   in   a   closet.

Set Up

1. Plug   in   power   cord   to   an   electrical   outlet.

2. Connect   Humidifier   (if   so   prescribed).

WARNING

DO   NOT   overfill   humidifier.

DO   NOT   reverse   the   oxygen   input   and   output   connections.

 

Water   from   the   humidifier   bottle   will   travel   through   the   cannula   back   to   the   patient.

NOTE:   For   this   procedure,   refer   to   FIGURE   6.1,   FIGURE   6.2

  on   page 19,   FIGURE   6.3

  on   page 20.

1. Remove   cap   from   bottle.

2. Fill   humidifier   with   distilled   water   to   the   level   indicated   by   the   manufacturer.

  Replace   the   humidifier   cap   and   securely   tighten.

 

Humidifier

Bottle with Cap

Humidifier

Bottle w/o Cap

Perfecto

2

™Series

FIGURE 6.1 Filling the Humidifier

18 Part No 1160877

SECTION 6—OPERATING INSTRUCTIONS

3. Insert   a   flathead   screwdriver   in   the   plate   groove   on   the   top   edge   of   the   filter   access   door   and   gently   pry   the   filter   access   door   off   (FIGURE   6.2).

4. Pull   up   and   remove   the   humidifier   bottle   adapter  

(FIGURE   6.2).

5. Replace   the   filter   access   door   and   install   the   air   scoop   onto   the   filter   access   door   by   inserting   the   six   tabs   on   the   air   scoop   into   the   six   slots   on   the   sides   of   the   filter   access   door.

Insert Flat

Head

Screwdriver

Here

Humidifier Bottle

Adapter

Filter Access

Door

Cabinet Filter

Air Scoop

Slots

Tabs

FIGURE 6.2 Humidifer Bottle Adapter

6. Attach   the   humidifier   bottle   adapter   to   the   humidifier   bottle   by   turning   the   wing   nut   on   the   humidifier   bottle   counterclockwise   until   it   is   securely   fastened.

  See   Detail  

“A”   in   FIGURE   6.3.

7. Place   the   humidifier   bottle/adapter   assembly   in   the   humidifier   compartment   on   the   concentrator.

  See   Detail  

“B”   in   FIGURE   6.3.

8. Attach   oxygen   tubing   from   the   humidifier   bottle/adapter   assembly   to   oxygen   outlet   connector   on   the   concentrator.

 

See   Detail   “B”   in   FIGURE   6.3.

9. Attach   the   cannula/patient   supply   tubing   to   the   humidifier   bottle   outlet.

  See   Detail   “B”   in   FIGURE   6.3.

Part No 1160877 19 Perfecto

2

™Series

SECTION 6—OPERATING INSTRUCTIONS

10. After   assembly,   ensure   that   oxygen   is   flowing   through   the   cannula.

DETAIL “A” DETAIL “B”

Humidifier

Bottle

Adapter

Oxygen

Tubing

Humidifier

Bottle

Humidifier

Bottle

Outlet

Oxygen

Outlet

Connector

Humidifier

Bottle

Humidifier

Compartment

FIGURE 6.3 Humidifier Compartment

Power Switch

NOTE:   For   this   procedure,   refer   to  

FIGURE   6.4.

1. Press   power   switch   to   On   position.

  All   the   panel   lights   and   the   audible   alarm   will   come   on   for   one   second,   indicating   that   the   unit   is   functioning   properly.

On/Off

(I/O)

Power

Switch

Perfecto

2

™Series 20

FIGURE 6.4 Power Switch

Part No 1160877

SECTION 6—OPERATING INSTRUCTIONS

Flowrate

NOTE:   For   this   procedure,   refer   to   FIGURE   6.5.

NOTE:   Take   care   not   to   set   the   flow   above   RED   ring.

  An   oxygen   flow   greater   than   5   L/min   will   decrease   the   oxygen   concentration.

1. Turn   the   flowrate   knob   to   the   setting   prescribed   by   your   physician   or   therapist.

WARNING

DO   NOT   change   the   L/min   setting   on   the   flowmeter   unless   a   change   has   been   prescribed   by   your   physician   or   therapist.

NOTE:   To   properly   read   the   flowmeter,   locate   the   prescribed   flowrate   line   on   the   flowmeter.

  Next,   turn   the   flow   knob   until   the   ball   rises   to   the   line.

  Now,   center   the   ball   on   the   L/min   line   prescribed.

 

2. If   the   flowrate   on   the   flowmeter   ever   falls   below   0.5

  L/min   for   more   than   about   one   minute,   the   LOW   FLOW   alarm   will   be   triggered.

  This   is   a   rapid   beeping   of   the   audible   alarm.

  Check   your   tubing   or   accessories   for   blocked   or   kinked   tubing   or   a   defective   humidifier   bottle.

  After   rated   flow   is   restored   to   more   than   0.5

  L/min,   the   LOW   FLOW   audible   alarm   will   go   off.

NOTE:   The   use   of   some   accessories   such   as   the   pediatric   flowstand   and   the   HomeFill   compressor   will   deactivate   the   Low   Flow   Alarm.

Flow Knob

Flowmeter

0.5 L/min

Ball

Front Panel

Part No 1160877

FIGURE 6.5 Flowrate

21 Perfecto

2

™Series

SECTION 6—OPERATING INSTRUCTIONS

SensO

2

Oxygen Purity Indicator - Models

IRC5PO2AW & IRC5PO2VAW

This   feature   monitors   the   purity   of   oxygen   generated   by   the   oxygen   concentrator.

  If   purity   falls   below   factory   preset   standards,   indicator   lights   on   the   control   panel   will   illuminate.

Initial Startup of the Concentrator

NOTE:   Concentrator   may   be   used   during   the   initial   start   warm ‐ up   time   (approximately   30   min.)   while   waiting   for   the   O

2

  purity   to   reach   maximum.

When   the   unit   is   turned   on,   the   GREEN   light   will   come   on  

(SYSTEM   OK/O

2

  greater   than   85%).

  After   five   minutes,   the   oxygen   sensor   will   be   operating   normally   and   will   control   the   indicator   lights   depending   on   oxygen   concentration   values.

 

Explanation of Oxygen Purity Indicator Lights -

Model IRC5PO2AW & IRC5PO2VAW

NOTE:   For   this   procedure,   refer   to   FIGURE   6.6

  on   page 23.

GREEN   light   (O

2

)  ‐  Normal   Operation.

YELLOW   light   ( )  ‐  Immediately   call   supplier.

  You   may   continue   to   use   the   concentrator   unless   instructed   otherwise   by   your   supplier.

  Be   certain   that   backup   oxygen   is   nearby.

RED   light   ( )  ‐  Total   unit   shutdown.

  Immediately   switch   to   backup   oxygen   supply   and   call   supplier.

GREEN   light  ‐  with   YELLOW   light   flashing  ‐  Immediately   call   supplier.

  Oxygen   sensor   malfunctioning;   you   may   continue   to   use   the   concentrator.

 

Perfecto

2

™Series 22 Part No 1160877

SECTION 6—OPERATING INSTRUCTIONS

OXYGEN PURITY

Auto Shut Down

Use Backup

Below

Normal

RED

Call Supplier

YELLOW

Normal O

2 GREEN

FIGURE 6.6 Explanation of Oxygen Purity Indicator Lights -

Model IRC5PO2AW & IRC5PO2VAW

Explanation of Indicator Lights - Model

IRC5PAW

NOTE:   For   this   procedure,   refer   to   FIGURE   6.7.

IRC5PAW Model Indicator Light Explanation

RED   light   ( )  ‐  Total   Unit   Shut ‐ Down.

  Immediately   switch   to   a   back ‐ up   oxygen   supply   and   call   supplier.

 

GREEN   light   (I/O)  ‐  On/Off.

  System   okay.

Auto Shut Down

Use Backup

RED

Call Supplier

Normal I/O GREEN

FIGURE 6.7 Explanation of Indicator Lights - Model IRC5PAW

Elapsed Time Meter (Hour Meter)

The   hour   meter   displays   the   cumulative   number   of   hours   the   unit   has   operated.

Part No 1160877 23 Perfecto

2

™Series

SECTION 7—MAINTENANCE

SECTION 7—MAINTENANCE

WARNING

The   Invacare   concentrators   are   specifically   designed   to   minimize   routine   preventive   maintenance.

  Only   qualified   personnel   should   perform   preventive   maintenance   on   the   concentrator.

 

Unplug   the   concentrator   when   cleaning.

  To   avoid   electrical   shock,   DO   NOT   remove   cabinet.

NOTE:   At   a   minimum,   preventive   maintenance   MUST   be   performed   according   to   the   maintenance   record   guidelines.

  In   places   with   high   dust   or   soot   levels,   maintenance   may   need   to   be   performed   more   often.

  Refer   to   Preventive   Maintenance   Record   on   page 26.

Routine Maintenance

The   following   routine   maintenance   should   be   performed   more   frequently.

  See   recommended   intervals   in   each   section   and  

Preventive   Maintenance   Record   on   page   26.

Cleaning the Cabinet Filter

CAUTION

DO   NOT   operate   the   concentrator   without   the   filter   installed.

NOTE:   For   this   procedure,   refer   to   FIGURE   7.1

  on   page 25.

 

NOTE:   There   is   one   cabinet   filter   located   on   the   back   of   the   cabinet.

1. Inspect   the   filter   monthly.

  Remove   the   filter   and   clean   as   needed.

NOTE:   Environmental   conditions   that   may   require   more   frequent   inspection   and   cleaning   of   the   filter   include,   but   are   not   limited   to:   high   dust,   air   pollutants,   etc.

Perfecto

2

™Series 24 Part No 1160877

SECTION 7—MAINTENANCE

2. Clean   the   cabinet   filter   with   a   vacuum   cleaner   or   wash   in   warm   soapy   water   and   rinse   thoroughly.

3. Dry   the   filter   thoroughly   before   reinstallation.

FIGURE 7.1 Cleaning the Cabinet Filter

Cleaning the Cabinet

1. Clean   the   cabinet   with   a   mild   household   cleaner   and   non ‐ abrasive   cloth   or   sponge.

Cleaning the Humidifier

NOTE:   To   clean   the   oxygen   humidifier,   follow   the   instructions   provided   by   the   manufacturer.

  If   none   are   provided,   follow   these  

STEPS:

1. Clean   the   humidifier   every   day.

2. Wash   it   in   soapy   water   and   rinse   it   with   a   solution   of   ten   parts   water   and   one   part   vinegar.

3. Rinse   thoroughly   with   hot   water   and   refill   with   distilled   water   to   the   level   shown   on   the   humidifier.

Part No 1160877 25 Perfecto

2

™Series

SECTION 7—MAINTENANCE

FIGURE 7.2 Preventive Maintenance Record

Perfecto

2

™Series 26 Part No 1160877

SECTION 8—TROUBLESHOOTING GUIDE

SECTION 8—TROUBLESHOOTING

GUIDE

SYMPTOM: PROBABLE CAUSE:

Alarm: Main Power Loss:

SOLUTION:

Short beeps, long pause

Concentrator not operating, power switch On.

Beep......Beep......

3. Tripped circuit breaker.

Alarm:

Continuous

1. Power cord not plugged in.

System Failure:

1. Unit overheating due to blocked air intake.

1. Insert plug into outlet.

2. No power at outlet.

2. Inspect house circuit breakers or fuses. If problem recurs, use a different outlet.

3. Push/reset circuit breaker.

If problem recurs, call Service

Provider.

1a. Remove and clean cabinet filters.

Concentrator not operating, Power

Switch On.

Beeeeeeeep....

2. Insufficient power at outlet.

3. Internal repairs required.

1b. Move oxygen concentrator at least three inches away from walls, draperies or furniture.

2. DO NOT use extension cords. Move to another electrical outlet or circuit.

3. Call Service Provider.

Part No 1160877 27 Perfecto

2

™Series

SECTION 8—TROUBLESHOOTING GUIDE

SYMPTOM: PROBABLE CAUSE:

YELLOW or RED light Illuminated.

SOLUTION:

1. Low oxygen purity.* 1. Clean or Replace filters.

Alarm:

Continuous

On RED light only.

* Only applies to

IRC5PO2AW &

IRC5PO2VAW models.

2. Kinked or blocked tubing, cannula or humidifier.*

3. Flowmeter set at 0.5

L/min *.

4. Unit overheating due to blocked air intake.

5. Insufficient power at outlet.

2. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for

60 seconds and then turn power back On.

3. Check flowmeter is set to

1.0 L/min or more. Refer to

Typical Product Parameters on page 13.

4a. Remove and clean cabinet filters.

4b. Move concentrator at least three inches from walls, draperies, and furniture.

5. DO NOT use extension cords. Move to another electrical outlet or circuit.

6. Call Service Provider.

GREEN light with

YELLOW Light

Flashing.

Alarm:

Rapid

Beep....Beep...

Beep...Beep

6. Internal repairs required.

1. Internal repairs required.

LOW FLOW ALARM

1. Kinked or blocked tubing, cannula or humidifier.

1. Call Service Provider.

1. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for

60 seconds and then turn power back ON.

2. Flowmeter set at 0.5

L/min.

2. Flowrates less than 1 L/min are not recommend.

Perfecto

2

™Series 28 Part No 1160877

SECTION 9—OPTIONAL ACCESSORIES

SECTION 9—OPTIONAL

ACCESSORIES

The   following   optional   accessories   are   also   available:  

• Standard   Adult   Nasal   Cannula,   2.1

  m   (7   feet) ‐  M3120

• Standard   Humidifier  ‐  M5120

• PreciseRX™   Pediatric   Humidifier/Flowmeter  

Accessory  ‐  IRCPF16AW

• HomeFill™   home   oxygen   compressor  ‐  IOH200AW

Part No 1160877 29 Perfecto

2

™Series

RECYCLING INFORMATION

RECYCLING INFORMATION

This   product   has   been   supplied   from   an   environmentally   aware   manufacturer   that   complies   with   the   Waste   Electrical   and   Electronic   Equipment   (WEEE)   Directive   2002/96/CE.

This   product   may   contain   substances   that   could   be   harmful   to   the   environment   if   disposed   of   in   places   (landfills)   that   are   not   appropriate   according   to   legislation.

The  ʹ crossed   out   wheelie   bin ʹ  symbol   is   placed   on   this   product   to   encourage   you   to   recycle   wherever   possible.

Please   be   environmentally   responsible   and   recycle   this   product   through   your   recycling   facility   at   its   end   of   life.

Perfecto

2

™Series 30 Part No 1160877

LIMITED WARRANTY

LIMITED WARRANTY

Belgium   &   Luxemburg:

Invacare   nv,   Autobaan   22,   B ‐ 8210  

Loppem

Tel:   (32)   (0)   50   83   10   10

Fax:   (32)   (0)   50   83   10   11 [email protected]

Danmark:  

Invacare   A/S,   Sdr.

  Ringvej   37,  

DK ‐ 2605   Brøndby

Tel:   (45)   (0)36   90   00   00

Fax:   (45)   (0)36   90   00   01 [email protected]

 

Deutschland:

Invacare   Aquatec   GmbH,  

Alemannenstraße   10,   D ‐ 88316   Isny

Tel:   (49)   (0)75   62   7   00   0

Fax:   (49)   (0)75   62   7   00   66 info@invacare ‐ aquatec.com

 

European   Distrbutor   Organisation:

Invacare,   Kleiststraße   49,   D ‐ 32457  

Porta   Westfalica  

Tel:   (49)   (0)57   31   754   540

Fax:   (49)   (0)57   31   754   541 [email protected]

España:

Invacare   SA,   c/Areny   s/n,   Polígon  

Industrial   de   Celrà,   E ‐ 17460   Celrà  

(Girona)

Tel:   (34)   (0)972   49   32   00

Fax:   (34)   (0)972   49   32   20 [email protected]

 

France:

Invacare   Poirier   SAS,   Route de   St  

Roch,   F ‐ 37230 Fondettes

Tel:   (33)   (0)2   47   62   64   66

Fax:   (33)   (0)2   47   42   12   24 [email protected]

 

Ireland:

Invacare   Ireland   Ltd,   Unit   5   Seatown  

Business   Campus,

Seatown   Road,   Swords,   County  

Dublin   –   Ireland

Tel:   (353)   1   810   7084

Fax:   (353)   1   810   7085 [email protected]

Italia:

Invacare   Mecc   San   s.r.l.,   Via dei   Pini  

62,   I ‐ 36016   Thiene   (VI)

Tel:   (39)   0445   38   00   59

Fax:   (39)   0445   38   00   34 [email protected]

 

Nederland:

Invacare   BV,   Celsiusstraat   46,  

NL ‐ 6716   BZ   Ede

Tel:   (31)   (0)318   695   757

Fax:   (31)   (0)318   695   758 [email protected]

  [email protected]

 

Norge:

Invacare   AS,   Grensesvingen 9,  

Postboks   6230,   Etterstad,   N ‐ 0603  

Oslo

Tel:   (47)   (0)22   57   95   00

Fax:   (47)   (0)22   57   95   01 [email protected]

  [email protected]

 

Õsterreich:

Invacare   Austria   GmbH,   Herzog  

Odilostrasse 101,   A ‐ 5310 Mondsee

Tel:   (43)   6232   5535   0

Fax:   (43)   6232   5535   4 info@invacare ‐ austria.com

 

Sverige   &   Suomi:

Invacare   AB,   Fagerstagatan 9,  

S ‐ 163 91 Spånga

Tel:   (46)   (0)8   761   70   90

Fax:   (46)   (0)8   761   81   08 [email protected]

[email protected]

 

Switzerland:

Invacare   AG,  

Benkenstrasse   260,  

CH ‐ 4108 Witterswil

Tel:   (41)   (0)61   487   70   80

Fax:   (41)   (0)61   487   70   81 [email protected]

Part No 1160877 31 Perfecto

2

™Series

United   Kingdom:

Invacare   Limited,   Pencoed  

Technology   Park,   Pencoed,  

Bridgend   CF35   5HZ

Tel:   (44)   (0)   1656   776200

Fax:   (44)   (0)   1656   776201

Customer   Services:

Tel:   (44)   (0)   1656   776222

Fax:   (44)   (0)   1656   776220

[email protected]

Portugal:

Invacare   Lda,  

Rua   Estrada   Velha,   949,   P ‐ 4465 ‐ 784  

Leça   do   Balio

Tel:   (351)   (0)225   1059   46/47

Fax:   (351)   (0)225   1057   39 [email protected]

 

Invacare Corporation

USA

One Invacare Way

Elyria, Ohio USA

44036-2125

440-329-6000

800-333-6900

Technical Services

440-329-6593

800-832-4707

EC REP www.invacare.com

Manufacturer:

Invacare Medical Equipment

(Suzhou) Co., Ltd.

Unit 2B, Export Processing

Zone

No. 200 Suhong Road SIP

(Suzhou Industrial Park)

Suzhou, Jiangsu, PRC

215021

Tel: 86-512-62586192

Fax: 86-512-62586072

EU Representative

Invacare International Sarl

Route de Cité Ouest 2

1196 Gland

Switzerland

Tel: +41 22 354 60 10

Fax: +41 22 354 60 11

© 2010 Invacare Corporation. All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.

Part No 1160877 Rev C - 09/10