Cambridge SSCSW-10AT Indicator Manual
Add to My manuals15 Pages
The CAMBRIDGE SSCSW-10AT is a versatile and feature-rich device designed to meet your weighing needs with precision and accuracy. With a capacity range of 1 to 999,999 pounds and a count-by feature for flexible division settings, it caters to a wide spectrum of applications. Its NTEP-approved model ensures compliance with legal for trade requirements, while the angle correction capability makes it ideal for lift truck scales.
advertisement
![Cambridge SSCSW-10AT Indicator Manual | Manualzz Cambridge SSCSW-10AT Indicator Manual | Manualzz](http://s3.manualzz.com/store/data/056118279_1-61d347861adaa2cb7b2dfa012937f3c0-360x466.png)
TABLE OF CONTENTS
Meter connections ------------------------------------------------------- 3
1.0 Operation ------------------------------------------------------ 4
1 .1 Key functions --------------------------------------------------- 4
1.2 Error Messages ----------------------------------------------------- 4
2 .0 Scale Procedure ------------------------------------------------------5
2 .1 Software Navigation Flowchart ---------------------------------- 5 & 6
2 .2 Navigation Keys ------------------------------------------------6
2 .3 Scale Menu Definitions -----------------------------------------7 & 8
2 .4 Numeric Entries ------------------------------------------------- 9
2.5 Set up Parameters ---------------------------------------------------- 9
3 .0 Calibration Procedures ---------------------------------------------- 1 0
3 .1 Calibration Menu Definitions ----------------------------------- 1 0
3 .2 Calibration --------------------------------------------------------- 1 1
3 .3 Linearity Correction ---------------------------------------------- 1 1
4 .0 Communications Setup ------------------------------------------------ 1 2
4 .1 Communications Menu Definitions ------------------------------ 1 2
5 .0 Testing Procedure ------------------------------------------------------ 1 3
5 .1 Testing Menu Definitions ---------------------------------------- 1 3
6 .0 Warr a nty ----------------------------------------------------------------- 1 4
7 .0 Assistance ----------------------------------------------------------------- 1 5
CAMBRIDGE SCALE WORKS, INC.
file:///C|/Manual/LFT%20TABLE%20OF%20CONTENTS.htm6/14/2006 10:50:10 AM
2 MANUAL P/N: 5999-1026-01
Meter Connections
To connect power to the
SS CSW-10 AT
Meter. First make sure the On/Off Switch on the Rear of the meter is in the
OFF
position. Connect the meter power cable to your
110VAC outlet.
ON / OFF
SWITCH
110VAC
POWER
CABLE
SIGNAL CONNECTION
+5V
GND
RXD
TXD
CAMBRIDGE SCALE WORKS, INC.
3
PIN # DESCRIPTION
A ------------ +EXC
B ------------ -EXC LOAD CELL
C ------------ +SIG
D ----------- -SIG
E ------------ +SEN
F ------------ -SEN
MANUAL P/N: 5999-1026-01
B ----------- -EXC
1.0
OPERATION
1.
1
Key Functions:
ZERO Brings the scale to a zero balance reading.
If the Zero key is pressed and held for 5 seconds the
Calibration zero value will be displayed.
GRS/NET
TARE
Toggles the display between Gross weight and Net weight.
This Key is also used to enter setup mode. Begin by pressing and holding this key until the Parameter ( Pxxx ) Event counter is displayed, then release. Immediately after COdE is displayed enter in sequence (within 5 sec) Tare , lb/kg
Print/Enter The display will indicate ScAlE .
, GRS/NET , and
-P xxx and C xxx are event counters that will increment each time one or more changes are made to the scale or Calibration Parameters.
Enters the Gross weight value into the Tare display and
switches to the Net display mode.
If the Tare key is pressed and held for 5 seconds the current Tare value will be displayed.
Toggles the display between pounds and kilograms.
lb/kg
PRINT Outputs the displayed weight data to the RS-232 Port.
Note: All keys are disabled when the scale is in motion or overload.
1.2 Error Messages
SCnEg
When the weight is more than 10 divisions negative from the zero calibration point.
The Scale is in an overload condition.
oLD bAt LO Will flash when the Battery voltage falls to 6.9VDC and will be displayed continuously when the voltage falls to 6.5VDC.
Err d More than 5 , 000 scale divisions have been selected in S1 Ntep and S1 angle mode.
More than 20,000 scale divisions have been selected in
S1 NO mode.
CAMBRIDGE SCALE WORKS, INC.
4 MANUAL P/N: 5999-1026-01
2
SCALE CALIBRATION
S1
Ntep
S2
CAPACITY
Ntep,no,angle
(Ntep)
1-999,999
(5000)
C1
Zero & Clear All
Points
C2
ZERO
SHOW RAW
COUNTS THEN 0
SHOW RAW
COUNTS THEN 0
S3
COUNT BY
1,2,5,/DEC
(1)
C3
SPAN POINT
SHOW ACTUAL
WEIGHT
S4
OVERLOAD
105% OF CAP
OR ENTERED
C4
SPAN POINT
S5
ZERO LIMIT
100%,1.9%
(100%)
C5
SPAN POINT
S6
FILTER
0 fastest to 7
slowest (3)
C6
SPAN POINT
S7
MOTION BAND
1 TO 99
Divisions (2)
S8
MOTION DELAY
0 to 99
UPDATES (4)
C7
SPAN POINT
END
CAMBRIDGE SCALE WORKS, INC.
2.0 SCALE PROCEDURE
2.1 SOFTWARE NAVIGATION FLOWCHART
RS-232
R1
BAUD RATE
1200,8,N,1-
19200,8,N,1?
R2
Output Format
R4
Receiver mode
WT ONLY,
G, N (GTN)
R3
Output Type
(Key) Polled,
Continuous
(Disabled)
Rec.,Remote
END
5
TEST
T1
VERSION
T2
DISPLAY
T3
BUTTONS
SOFTWARE
VERSION
LIGHT
SEGMENTS
b1-b4,
b5 EXITS
T4
A TO D
T5
SERIAL
T6
Cal Factors ?
A/D RAW
COUNTS
Jumper Rx,Tx
Loop Back
T7
DEFAULT
YES, NO
END
END
MANUAL P/N: 5999-1026-01
S1
Ntep
S2
CAPACITY yes,no,angle
(yes)
1-999,999
(5000)
S3
COUNT BY
1,2,5,/DEC
(1)
S4
OVERLOAD
105% OF CAP
OR ENTERED
S5
ZERO LIMIT
100%,1.9%
(100%)
S6
FILTER
ADU Digital &
Bandwidth Filter
S7
MOTION BAND
1 TO 99
Divisions (2)
S8
MOTION DELAY
0 to 99
UPDATES (4)
S9
Motion Display
Normal, Blank,
Dashes (Normal)
2.1 SOFTWARE NAVIGATION FLOWCHART (CONTINUED)
S10
ZERO BAND
1-99
DIVISIONS(2)
S11
ZERO DELAY
0-99
UPDATES(2)
S12
Zero Tracking
S13
Tracking DELAY
0 = .5 Div.
(1) = 1 Div.
3 = 3 Div.
1-99
1/4secinc.
(0)
S14
Display lb/kg lb ,kg, lb/kg
(lb/kg)
S15
SCALE ID
1-99 (1)
ID IN TX
S16
BRIGHTNESS
1-15 (15)
S17
SLEEP MODE
0 to 30 min blanks dsp (0)
END
2.2 NAVIGATION KEYS
During setup you will be required to make numeric entries. (Ex: Capacity, Zero Band, etc...)
The following table outlines the keys used to perform these entries along with their function.
ZERO-------KEY IS USED TO NAVIGATE UP.
GRS/NET---KEY IS USED TO NAVIGATE DOWN.
TARE--------KEY IS USED TO NAVIGATE LEFT.
lb/kg---------KEY IS USED TO NAVIGATE RIGHT.
PRINT-------KEY IS USED TO ENTER DATA AND RETURNS.
CAMBRIDGE SCALE WORKS, INC.
6 MANUAL P/N: 5999-1026-01
2.3
Scale Menu Definitions:
S1 NTEP (Ntep): Maximum divisions limited to 5 , 000.
scale negative message is displayed if the gross weight goes more than 10 divisions below zero. If a capacity and count by of more than 5 , 000 divisions is selected ERR d will be displayed and you will be returned to S2 to select a new capacity or count by.
Angle: Enables angle correction for Legal For Trade lift truck scales.
No: 20,000 maximum division limits and no scale negative tests.
S2 Capacity
S3 Count By
1 to 950,000 pounds (5
.001, .01, .1, (1), 10, 100,
.002, .02, .2, 2, 20, 200
.005, .05, .5, 5, 50, 500
, 000).
S4 Over load (105%) of scale capacity or user entered value.
S5 Zero Limit (100%) or 1.9% of scale capacity.
S6 Filter 0 - 7 (3) 0 is the fastest response or least filtering and 7 is the slowest response or most filtering
S7 Motion Band 1 to 99 Divisions (2) the weight display must be stable within the selected number of updates for the motion indicator to be turned off.
S8 Motion Delay 0 to 99 updates (4) the weight display must be within the motion band for the selected number of updates for the motion indicator to be turned off.
S9 Motion (Normal): When the scale is in motion the motion LED will
Display light.
Blank: When the scale is in motion the display will be blanked out.
Dashes: When the scale is in motion the display will show all Dashes.
S10 Zero Band 1 to 99 divisions (2) the weight display must return to Zero within the selected number of divisions to to be considered Zero.
CAMBRIDGE SCALE WORKS, INC.
7 MANUAL P/N: 5999-1026-01
2.3
Scale Menu Definitions Continued:
S11 Zero Delay 0 to 99 updates (4) the weight display must be within the zero band for the selected number of updates to be considered Zero.
S12 Zero 0 .5 Divisons
Tracking (1) 1 Division
3 3 Divisions
S13 Tracking Delay (0) to 99 Updates. the time that the display must be within the allowed graduations before it will automatically Zero.
(0) Disables Zero Tracking
S14 Lb/Kg (Lb/KG): Allows the indicator to be switched between pounds and kilograms by pressing the lb/kg key.
LB: This sets the display to Pounds only.
Kg: This sets the display to Kilograms only.
S15 Scale ID (1) to 99 scale ID used in RF link output.
S16 Brightness 0 to (15) adjusts the LED display intensity 15 is the brightest.
S17 Sleep Mode (0) to 30 minutes. The display will turn off after the the set amount of time elapses with no scale activity.
END Exits back to the main Menu.
( ) indicates Factory Set defaults.
CAMBRIDGE SCALE WORKS, INC.
8 MANUAL P/N: 5999-1026-01
1
2.4 Numeric Entries
When entering a numeric value, f irst press and release the lb/kg key to move right into the menu where the numeric value will be entered. Then press and release the Zero key, the first digit in the value will flash. Press and release the
ZERO and GRS/NET keys to increase or decrease the digits value. Press and release the Tare key to move to the next digit. Repeat the steps above to adjust the digits value. Repeat all steps until the numeric value is correct, then press and release the PRINT key to enter the data. The display will return to the menu.
2.5 Set Up Parameters
To begin press and hold the GRS/NET key until the Parameter ( Pxxx ), Event counter is displayed, then release. Immediat e ly after COdE is displayed, (within 5 sec.) enter in sequence indicate ScAlE .
Tare , l b/kg , GRS/NET , and Print/Enter , The display will
-P xxx and C xxx are event counters that will increment each time one or more changes are made to the Scale or Calibration Parameters.
With the display indicating ScAlE , press the GRS/NET key to move down. This allows the operator to change any of the scale parameter S1 thru S1 7 . For example
Press the GRS/NET key to move down until S2 is displayed, S2 is used to set the capacity of the scale. Press the lb/kg key to move right into the S2 Parameter. The current capacity will be displayed. Press the ZERO key, the first digit will flash. Enter the new capacity using the steps described in the above section 2.4. When the capacity is correct, press the Print key to enter the value. The display will return to S2 .
Entering the Count By. Press the GRS/NET to move down, S3 will be displayed.
Press the lb/kg key to move right into this parameter. The current "Count by" will be displayed. Press the ZERO and GRS/NET keys to adjust the divisions. Press the TARE and lb/kg keys to adjust the decimal point. Press the PRINT key, when finished, t o enter the data. The display will return to S3 . Exit Set up Mode By pressing the TARE key, End will be displayed, then press PRINT/ENTER .
CAMBRIDGE SCALE WORKS, INC.
9 MANUAL P/N: 5999-1026-01
3.0 CALIBRATION PROCEDURES
3.1
Calibration
Menu Definitions
:
C
1 Zero All Raw counts, (pitch and roll if in angle mode) will be displayed.
When ZERO is pressed an analog Zero is done and all calibration span points will be cleared.
If in angle mode the pitch and roll offsets will also be zeroed.
C
2 Zero Zeroed Raw counts, (pitch and roll if in angle mode will be displayed. When ZERO is pressed an analog zero is done and all calibration span points will NOT be cleared.
If in angle mode the pitch and roll offsets will also be zeroed.
C
3 Span Point The last Calibration weight will be displayed then the
actual weight on the scale will be displayed.
If you do not wish to change the span point, press the
TARE key to exit without making any changes.
If the displayed weight does not match the known test weight, press the ZERO key to enter the correct weight.
Use the steps d e scribed in section 1.4 for numeric entry.
When the weight is correct press the Print key to enter the new value.
C4 to C7 Span points C4 to C7 are for linearity correction, they can be used in order and in any quantity or not at all if no correction is
necessary.
You may also return to C4 to C7 later and add a new
correction point without affecting any original calibration points.
The last calibration weight will be displayed then the actual weight on the scale will be displayed. If no calibration weight has been entered at this span point "notset" will be displayed then the actual weight on the scale is displayed.
If the displayed weight does not match the known test weight, follow the steps described for C3 Span point on adjusting the weight and entering the value.
END Exits back to the main Menu.
CAMBRIDGE SCALE WORKS, INC.
10 MANUAL P/N: 5999-1026-01
3.2 Calibration
Press and hold the GRS/NET key as d e scribed p reviously in section 2.5. ScAlE will be displayed. Press the lb/kg key to move right, CALib will be displayed. Press the GRS/NET key to move down, C1 will be displayed. Press the lb/kg key to move right, The Raw counts will be displayed. With no weight on the scale and the scale level, press the ZERO button, “0” will be displayed. Press the PRINT key to enter the data. “0” is now entered and the display returns to C1 .
Note: With the scale completely level use C1 or C2 to zero. There is no need to use both C1 and C2.
Press GRS/NET key to move down, C3 will be displayed. Press the lb/kg key to move to the right, the last calibrated weight will flash then the current weight on the scale is displayed. Place a known test weight on the scale with the forks level. Press the ZERO key, the first digit of the weight will flash. Use the ZERO and GRS/NET keys to increase or decrease the digits value. Press the TARE key to move left, the next digit will flash. Repeat the steps until the correct weight is entered. Press the
PRINT key to record the data. The display will return to C3 .
Press the GRS/NET key until End is displayed, then Press PRINT . Calibration is now complete.
3.3 Linearity Correction
If Linearity Correction is needed, Press the GRS/NET key (from the calibration menu) to move down to C4 . Press the lb/kg key to move right, the last calibrated weight will flash or “ notset ” will flash if this point has not been previously set. Next the current weight on the scale will be displayed. Place a known test weight on the scale with the forks level. Press the ZERO Key for the first digit of the displayed weight to flash. Enter the weight as described in section 1.4, then press the PRINT key to record the data. Continue these steps for C5 , C6 and C7 .
Linearity Correction points ( C4 C7 ) can be used in any order and in any quantity or not at all if no correction is necessary. After calibration is complete you may also return to these correction points and make changes to its value without affecting any of the original calibration points.
CAMBRIDGE SCALE WORKS, INC.
11 MANUAL P/N: 5999-1026-01
4.0 COMMUNICATIONS SETUP
4.1
Communications
Menu Definitions
:
R
1 Baud Rate 1200 to 115200 baud (9600), 8, n, 1
R
2 Output 0 - (Gross, Tare, Net) Gross only, Net only
Format 1 - Weight only
2 - Net only
3 - Gross only
R
3 Output Type 0 - Output on command, standard print.
Output as selected by R2 Output Format.
If "Q" is received on the serial port the scale will output the same as if the PRINT key were Pressed.
The same holds true for Z = Zero
U = lb/kg
D = GRS/NET
T = TARE
1 - Slave Display Output (numeric only) continuous
Stx, Six ASCII Characters (indicated weight), CR, LF
2 - Slave Display Output (alphanumeric) continuous
Stx GR or NT or TR, six ASCII Characters (indicated
weight), lb or kg, CR, LF.
3 - RF Link Output.
4 - Used for QSI terminal.
R4 Receiver 0 - Disabled normal scale mode.
1 - Standard Receiver.
Receives R3.3 RF link output string displaying data as
it appears on the scale. All keys are disabled except the
PRINT key. Receive "B" will print the same as if the button were pressed.
2 - Remote control.
Receives R3.3 RF link output string displaying data as it appears on the scale and allows full control of all scale meter functions.
END Exits back to the main Menu.
( ) indicates Factory Default setting
CAMBRIDGE SCALE WORKS, INC.
12 MANUAL P/N: 5999-1026-01
5.0 TESTING PROCEDURES
5.1 Testing
Menu Definitions
:
T
1 Version Displays Software Version.
T
2 Display Lights all display segments and indicating LED s
T
3 Buttons Press the Zero key, b1 will be displayed.
Press the GRS/NET key, b2 will be displayed.
Press the Tare key, b3 will be displayed.
Press the lb/kg key, b4 will be displayed.
Pressing the print key will exit back to the menu T3.
T4 A to D
T5
T6
T7
Serial
Setup data
Default
END
Displays Raw counts where a 1mV/V signal from the scale will display 25,000 counts.
When in angle mode Down GRS/NET will cycle Pitch, Roll and Raw counts.
Serial communications can be verified by connecting pins
2 and 3 on the serial port. A single character will be echoed
and pass or fail will be displayed.
Setup data will be sent out on the printer port.
Resets the meter back to factory defaults clearing all calibration and setup data. "r you sure?" will be displayed then press the Tare key to exit without Defaulting, or press the Print key to reset the meter to factory default.
Exits back to the main Menu.
CAMBRIDGE SCALE WORKS, INC.
13 MANUAL P/N: 5999-1026-01
6.0 WARRANTY
CAMBRIDGE warrants the SS CSW-10AT to be free of defects in workmanship and/or materials for 12 months from the date of shipment . This warranty of workmanship and/or materials, is valid, if in the opinion of CAMBRIDGE , the equipment has not been mechanically, environmentally, or electrically abused.
This warranty is limited, at the option of
CAMBRIDGE
, to repair, replace or an appropriate credit adjustment not to exceed the original equipment sale price paid to
CAMBRIDGE . CAMBRIDGE assumes no liability in connection with the sales of its products beyond that stated above.
Warranty replacement parts and/or repair services are performed at the factory in Cumberland, Maryland or by an authorized Service Group approved by
CAMBRIDGE .
Warranty does not include travel expense if a factory technician is requested to perform repairs at a location other than the factory.
It the user's responsibility to follow the proper set-up, calibration and operating procedures of the SSCSW-10
AT as d e scribed in this manual. If the operator has difficulty using their
SSCSW-10
AT indicator properly, please contact
CAMBRIDGE at 1-301-724-4082. Any one of our Technicians will be happy to work with the user via telephone.
Thank You!
CAMBRIDGE SCALE WORKS, INC.
14 MANUAL P/N: 5999-1026-01
SW-10-LFT to be free of defects in workmanship and/or materials for 12 months from the date of delivery. This warranty of
7.0 Assistance
If at any time you require assistance with you
r
Model
:
S SCSW-10 AT
Meter please contact us at:
CAMBRIDGE SCALE WORKS, INC.
115
West Mary Street
Cumberland, MD 21502
Phone:
(301) 724-4082
Fax: (301) 724-4964
CAMBRIDGE SCALE WORKS, INC.
15 MANUAL P/N: 5999-1026-01
advertisement
Related manuals
advertisement