Luminostics clipcovid Rapid Antigen Test Instructions for use
تبلیغات
تبلیغات
PACKAGE INSERT (INSTRUCTIONS FOR USE)
For Use Under Emergency Use Authorization (EUA) Only
For Use With Anterior Nasal Swab Specimens
For in vitro Use Only
Rx Only
INTENDED USE
The Clip COVID Rapid Antigen Test comprises the Clip Analyzer and the Clip COVID Rapid Antigen
Test Kit. The Clip COVID Rapid Antigen Test is a lateral flow immunoluminescent assay intended for the qualitative detection of nucleocapsid protein antigen from SARS-CoV-2 directly from anterior nasal swab specimens collected from individuals who are suspected of COVID-19 by their healthcare provider within the first five days of onset of symptoms. Testing is limited to laboratories certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. §263a, that meet the requirements to perform high, moderate, or waived complexity tests. This test is authorized for use at the Point of Care (POC), i.e., in patient care settings operating under a CLIA Certificate of
Waiver, Certificate of Compliance, or Certificate of Accreditation.
The Clip COVID Rapid Antigen Test does not differentiate between SARS-CoV and SARS-CoV-2.
Results are for the identification of SARS-CoV-2 nucleocapsid protein antigen. Antigen is generally detectable in anterior nasal swab specimens during the acute phase of infection. Positive results indicate the presence of viral antigens, but clinical correlation with patient history and other diagnostic information is necessary to determine infection status. Positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.
Laboratories within the United States and its territories are required to report all results to the appropriate public health authorities.
Negative results from patients with symptom onset beyond five days should be treated as presumptive and confirmation with a molecular assay, if necessary, for patient management, may be performed.
Negative results do not rule out SARS-CoV-2 infection and should not be used as the sole basis for treatment or patient management decisions, including infection control decisions. Negative results should be considered in the context of a patient’s recent exposures, history and the presence of clinical signs and symptoms consistent with COVID-19, and confirmed with a molecular assay, if necessary for patient management.
Luminostics, Inc. | Page 1 of 16
The Clip COVID Rapid Antigen Test is intended for use by healthcare professionals or individuals
SUMMARY AND EXPLANATION OF THE TEST
Coronaviruses are a family of RNA viruses; a subset of coronaviruses cause illness in animals or humans.
SARS-CoV-2 is a coronavirus that can cause mild to severe respiratory illness and has spread globally beginning in late 2019. The Clip COVID Rapid Antigen Test is a rapid test for the qualitative detection and diagnosis of SARS-CoV-2 directly from nasal swabs. The Clip COVID Rapid Antigen Test Kit, along
TEST PRINCIPLE
The Clip COVID Rapid Antigen Test employs persistent luminescence immunoassay technology in a sandwich lateral flow assay design to detect SARS-CoV-2 nucleocapsid protein from anterior nasal swab specimens. The patient’s nasal sample is placed in the Extraction Tube, during which time the virus particles in the sample are disrupted, releasing viral nucleoproteins. The extracted sample is dispensed into the Cartridge’s sample well from where it migrates through a lateral flow test strip containing various chemical environments. If SARS-CoV-2 viral antigen is present, it will be trapped in a specific location and be labeled by a persistent luminescent reporter nanoparticle. The Clip algorithms are used to display objective test results (Positive, Negative, or Invalid) on the screen.
Analyzer then measures a luminescence signal from the test strip following which method-specific
MATERIALS SUPPLIED
● Cartridges (25), individually packaged in foil pouches and containing lateral flow test strips
● Extraction Tubes (25 unitized tubes), each containing 500 µL of assay reagent
● Dropper Tips (25)
● Sterile Nasal Swabs (25)
● Positive Control Swab (1), non-infectious recombinant SARS-CoV-2 nucleocapsid antigen
● Negative Control Swab (1), blank nasal swab
● dried onto a swab
Package Insert (1)
Materials required but not supplied:
● Clip Analyzer
● Timer, clock, or watch
WARNINGS AND PRECAUTIONS
● For
●
in vitro
diagnostic use.
For prescription use only
Luminostics, Inc. | Page 2 of 16
●
FDA under an Emergency Use Authorization (EUA) for use by laboratories certified under the
CLIA, 42 U.S.C. §263a, that meet the requirements to perform moderate, high or waived complexity tests.
● This test has been authorized only for the detection of proteins from SARS-CoV-2, not for any other viruses or pathogens.
● The emergency use of this test is only authorized for the duration of the declaration that circumstances exist justifying the authorization of emergency use of in vitro diagnostic tests for detection and/or diagnosis of COVID-19 under Section 564(b)(1) of the Federal Food, Drug and
Cosmetic Act, 21 U.S.C. § 360bbb-3(b)(1), unless the declaration is terminated or authorization is revoked sooner.
● Do not use the Test Kit contents beyond the expiration date printed on the outside of the box.
● Do not reuse a used Test Kit or any elements therein.
● The Clip COVID Rapid Antigen Test is designed for counter top operation.
● The Clip COVID Rapid Antigen Test is not designed to withstand moisture, extreme humidity, or extreme temperatures. Use under these conditions may cause false positive or false negative results.
● Do not open the foil pouch of the Cartridge and expose it to the ambient environment until the
Cartridge is ready for immediate use (within 30 seconds of opening foil pouch). Premature exposure to ambient conditions may cause false positive, false negative, or invalid results.
● Discard and do not use any damaged or dropped Cartridge or material. This may result in a cracked or misaligned Cartridge which may cause false positive, false negative, or invalid results.
● The reagent in the Extraction Tube contains sodium azide. If the solution contacts the skin or eye, flush with copious amounts of water.
● To obtain accurate results, the Package Insert must be followed.
● Inadequate or inappropriate sample collection, storage, and transport may yield false test
● Always operate the Clip Analyzer and use other components of the Clip COVID Rapid Antigen
● results. cause false positive, false negative, or invalid results.
● As the test is based on a luminescent immunoassay, no visible results will form on the test strip.
●
The Clip Analyzer must be used for result interpretation.
Test on a surface that is level, dry, and not in direct sunlight. Use under these conditions may
Do not move or adjust the Clip Analyzer or remove the Cartridge while there is a test in
Sample collection and handling procedures require specific training and guidance. Please read
● When collecting a nasal swab sample, use the Nasal Swab supplied in the kit.
● The Clip Analyzer must be used for result interpretation.
● progress. Doing so may cause an invalid result. the entirety of this package insert and the user manual prior to executing a test.
Wear suitable protective clothing, gloves, and eye/face protection when handling the contents of this kit.
● To reduce the risk of biohazard:
Luminostics, Inc. | Page 3 of 16
○ Use appropriate precautions in the collection, handling, storage, and disposal of patient samples and used kit contents.
○ Use of Nitrile, Latex (or equivalent) gloves is recommended when handling patient samples. Change gloves between patients.
○ Dispose of used specimens and test kit components in accordance with Federal, State, and Local requirements.
○ Treat specimens and patient samples as well as used test kit components as potentially biohazardous materials.
○ Ensure the Analyzer is cleaned per the Cleaning Guidelines in this Package Insert and the Analyzer User Manual.
○ Wash hands thoroughly after handling.
● The Clip COVID Rapid Antigen Test contains small parts that may be dangerous if swallowed.
● The product has not been tested for EMI compatibility with implantable cardioverterdefibrillators (ICDs) or pacemakers. Do not use the Clip Analyzer if you have an ICD or pacemaker.
STORAGE AND STABILITY
Store the Clip COVID Rapid Antigen Test at room temperature, 59°F to 86°F (15°C to 30°C), out of direct sunlight. Kit contents are stable until the expiration date printed on the outer box. Do not refrigerate or freeze.
QUALITY CONTROL
There are two types of Quality Controls for the Clip COVID Rapid Antigen Test: built-in procedural control features and external positive/negative controls.
Procedural Control
The Clip COVID Rapid Antigen Test contains a built-in procedural control feature. The procedural control is interpreted by Clip Analyzer after the run time of the test. If the test does not run correctly, the test with a new patient sample and a new Cartridge, Extraction Tube, and Dropper Tip.
External Positive and Negative Controls
External Controls may also be used to demonstrate that the reagents and assay procedure perform properly. A Positive Control Swab and Negative Control Swab are included as External Controls.
Luminostics recommends that Positive and Negative Control Swabs be run once for each untrained operator and as deemed additionally necessary by your internal quality control procedures and in accordance with Local, State and Federal regulations or accreditation requirements. Patient tests should not be performed if the test fails to detect Positive and/or Negative Control Swabs accurately; the control tests should be repeated or Luminostics Support should be contacted at [email protected].
Luminostics, Inc. | Page 4 of 16
SPECIMEN COLLECTION
Use the nasal swab supplied in the kit. Inadequate specimen collection may yield erroneous results.
To collect an anterior nasal swab sample using the swab supplied in the kit:
1. Insert the tip of the swab in the vertical position into one nostril until there is gentle resistance at the level of the turbinates (less than one inch into the nostril). The entire tip of the swab (usually ½ to ¾ of an inch) should be placed inside the nose, and the side of the resistance at the level of the turbinates (less than one inch into the nostril). Keep the swab in
other
nostril until there is gentle place and rotate FIVE (5) times against the nasal wall (five complete rotations) and gently swab tip should be rubbed with moderate pressure against as much of the wall of the anterior nares region as possible, moving the tip through a large circular path inside the nose.
2. Keep the swab in place and rotate FIVE (5) times against the nasal wall (five complete rotations) and gently remove from the nose. This should take approximately 10-15 seconds per nostril.
3. Gently insert the swab in the vertical position into the
CAUTION! Simply twirling the swab against one part of the inside of the nose or leaving the swab in the nose for 10-15 seconds, is not proper technique and may result in an insufficient
SPECIMEN TRANSPORT AND STORAGE
For best performance, nasal swab specimens should be tested as soon as possible (within 30 minutes extraction buffer, and storage or retesting from specimens in extraction buffer is not recommended. banked (frozen) samples. stored dry at room temperature) after collection. No stability data is available for specimens stored in
The Clip COVID Rapid Antigen Test kit has not been tested for use with viral transport media or
TEST PROCEDURE
1. Place the Clip Analyzer on a table or counter top and power it on as instructed in the
Analyzer manual. The Analyzer is portable and can be moved to a suitable location for provides adequate space for the Clip Analyzer. There must be space to access the Clip
Analyzer port for insertion of the Cartridge. We recommend that you keep the Analyzer
2. Remove a Cartridge from its foil pouch after using the tear notch to open the pouch. testing. Ensure the surface is stable, level, dry and free of obstructions. Ensure the bench plugged in to a power outlet using the provided charging cord during operation/testing.
Luminostics, Inc. | Page 5 of 16
3. Load the Cartridge into the Analyzer by pushing the Cartridge into the Cartridge port until you hear a click. If you don’t hear a click, continue pushing the Cartridge until you can’t push it further.
5. Enter sample ID and select the Test Kit lot number from the drop-down menu. Touch
‘Continue’.
6. Confirm the adapter ID by matching the label on the side of the adapter to the unit ID on the screen. Touch ‘Confirm’ if the ID on screen matches the ID on the side of the Analyzer.
Luminostics, Inc. | Page 6 of 16
7. Insert the anterior nasal swab collected from the patient all the way into the bottom
Extraction Tube and rotate the swab 3 times against the bottom of the tube.
8. Leave the swab in the buffer in the Extraction Tube for 60 seconds.
9. Squeeze the center of the Extraction Tube and remove the swab while keeping the center of the tube squeezed. Dispose of swab in a biohazard waste stream.
11. Dispense the entirety of the contents of the Extraction Tube into the sample well of the
Cartridge by turning it upside down and squeezing it. Holding the tube vertically directly above the sample port will minimize spillage.
Luminostics, Inc. | Page 7 of 16
12. The Analyzer will automatically begin analysis 30-45 seconds after sample addition, transitioning to the ‘Analysis in Progress’ screen. A “positive”, “negative”, or “invalid” result will display in 30 minutes. Do not touch or remove the cartridge or disturb the Analyzer until result in an erroneous or invalid result.
RESULT INTERPRETATION
When the test is complete, the result will be displayed on the Analyzer screen. The result of the lateral flow test cannot be seen with the naked eye. The Analyzer screen will display results, individually providing a positive or negative result for SARS-CoV-2. If the result is Invalid, retest with a new patient
Positive Result Negative Result Invalid Result
Luminostics, Inc. | Page 8 of 16
This display shows a valid positive result for SARS-CoV-2.
This display shows a valid negative result for SARS-CoV-
2.
This display shows an invalid result.
LIMITATIONS
● The contents of this kit are to be used for the qualitative detection of SARS-CoV-2 antigens from an anterior nasal swab.
● This test detects both viable (live) and non-viable SARS-CoV-2. Test performance depends on the amount of virus (antigen) in the sample and may or may not correlate with viral culture results performed on the same sample.
● A negative test result may occur if the level of antigen in a sample is below the detection limit of the test or if the sample was collected improperly.
● Failure to follow the Test Procedure may adversely affect test performance and/or invalidate the test result.
● The test has not been validated for use with viral transport media (VTM) or universal transport media (UTM). Usage of the test with samples prepared using VTM or UTM may cause false positive, false negative, or invalid results.
● The test has been validated for use in temperatures ranging from 15 ° C-30 ° C. The test has not been validated for use in temperature ranges outside of these conditions and usage outside of the validated range of conditions may result in false positive results or false negative results.
● Test results must be evaluated in conjunction with other clinical data available to the physician.
● Positive test results do not rule out co-infections with other pathogens.
● Positive test results do not differentiate between SARS-CoV and SARS-CoV-2.
● Negative test results are not intended to rule in other non-SARS viral or bacterial infections.
● Negative results, from patients with symptom onset beyond five days, should be treated as presumptive and confirmation with an FDA authorized molecular assay, if necessary, may be performed for clinical management.
Luminostics, Inc. | Page 9 of 16
● If the differentiation of specific SARS viruses and strains is needed, additional testing, in consultation with state or local public health departments, is required.
CONDITIONS OF AUTHORIZATION FOR THE LABORATORY
The Clip COVID Rapid Antigen Test Letter of Authorization
, along with the authorized Fact Sheet for Healthcare Providers, the authorized Fact Sheet for Patients, and authorized labeling are available on the FDA website: https://www.fda.gov/medical-devices/emergency-situations-medical-devices/emergency-useauthorizations#covid19ivd.
However, to assist clinical laboratories using the Clip COVID Rapid Antigen Test (“your product” in the conditions below), the relevant Conditions of Authorization are listed below:
● Authorized laboratories using your product will include with test result reports, all authorized
Fact Sheets. Under exigent circumstances, other appropriate methods for disseminating these
Fact Sheets may be used, which may include mass media.
● Authorized laboratories using your product will use your product as outlined in the authorized labeling, e.g, “Clip COVID Rapid Antigen Test Package Insert (Instructions for Use” and “User
Manual-Clip Analyzer.” Deviations from the authorized procedures, including the authorized instruments, authorized clinical specimen types, authorized control materials, authorized other ancillary reagents and authorized materials required to use your product are not permitted.
● Authorized laboratories that receive your product will notify the relevant public health authorities of their intent to run your product prior to initiating testing.
● Authorized laboratories using your product will have a process in place for reporting test results to healthcare providers and relevant public health authorities, as appropriate.
● Authorized laboratories will collect information on the performance of your product and report to DMD/OHT7-OIR/OPEQ/CDRH (via email: [email protected]) and
Luminostics, Inc. (via email: [email protected], any suspected occurrence of false positive or false negative results and significant deviations from the established performance characteristics of your product of which they become aware.
● All operators using your product must be appropriately trained in performing and interpreting the results of your product, use appropriate personal protective equipment when handling this kit, and use your product in accordance with the authorized labeling.
● Luminostics, Inc., authorized distributors, and authorized laboratories and patient care settings
1 The letter of authorization refers to, “Laboratories certified under the Clinical Laboratory
Improvement Amendments of 1988 (CLIA), 42 U.S.C. §263a, that meet the requirements to perform high, moderate, or waived complexity tests. This test is authorized for use at the Point of Care (POC), i.e., in patient care settings operating under a CLIA Certificate of Waiver, Certificate of Compliance, or Certificate of Accreditation.” as “authorized laboratories.”
Luminostics, Inc. | Page 10 of 16
otherwise notified by FDA. Such records will be made available to FDA for inspection upon request.
CLINICAL PERFORMANCE
Patient Demographics
Patient demographics (age, elapsed time from date of on-set) are available for the one hundred sixtysix (166) samples used in the study. The specimen positivity breakdown based on age of the patient:
Age Range
Under 5 years
6 to 21 Years
22 to 59 years*
60 years and older
Total # of Enrolled
Subjects
0
11
147
8
Total # Positive by RT-
PCR
0
1
31
0
*one of the samples in this age range was negative on the Clip COVID Rapid Antigen Test but positive by RT-PCR
All patients enrolled in the study were symptomatic and provided at least one nasal and one nasopharyngeal swab. At both sites, one nasal swab was tested directly in the Clip COVID Rapid
Antigen Test, within 30 minutes of collection, according to product instructions. Nasopharyngeal swabs were eluted in viral transport media (VTM) and immediately frozen before being batched and shipped to a central laboratory for RT-PCR testing on an EUA-authorized assay that includes a solidphase RNA extraction step. The performance of the Clip COVID Rapid Antigen Test was established within 5 days of symptom onset. by testing 166 nasal swabs from individual symptomatic patients who were enrolled into the study
Comparator RT-PCR Assay
Clip COVID Rapid Antigen
Test by Luminostics, Inc.
Positive Negative Total
Luminostics, Inc. | Page 11 of 16
Positive
31 0 31
Negative
Total
1
32
134
134
Positive Percent Agreement: 96.9% (
95% CI:
83.8% - 99.9%)
Negative Percent Agreement: 100% (
95% CI:
97.3% - 100%)
135
166
The specimen positivity based on days post onset:
Days Post Symptom
Onset
# Specimens Tested from Unique People
0 23
3
4
1*
2
5
36
56
27
14
10
# Positive Specimens by
RT-PCR
0
8
6
5
12
1
% Positive
0
13.9%
21.4%
29.6%
42.9%
10%
*One specimen was Clip COVID Rapid Antigen Test Negative and Positive by Reference Extracted
PCR.
ANALYTICAL PERFORMANCE
Limit of Detection
The Limit of Detection (LoD) of the Clip COVID Rapid Antigen Test was determined using limiting dilutions of gamma-irradiated SARS-CoV-2 (BEI Resources NR-52287). NR-52287 is a preparation of
SARS-Related Coronavirus 2 (SARS-CoV-2), isolate USA-WA1/2020, that has been gamma-irradiated
(5x10 6 RADs) on dry ice, followed by sonication. The material was supplied frozen at a concentration of 2.8 x10 5 TCID
50 were eluted, combined, and mixed thoroughly to create a human nasal swab extract clinical matrix per mL. The study to determine the Clip COVID Rapid Antigen Test’s LoD was designed to reflect the assay when using direct swabs. Presumed negative natural nasal swab specimens
Luminostics, Inc. | Page 12 of 16
swab was spiked with 50 µL of the virus dilution in pooled nasal swab extract. The spiked swab was processed on the Clip COVID Rapid Antigen Test according to the package insert. The LoD was determined in two steps:
LoD Screening
Five (5) dilutions of the gamma irradiated virus were made in pooled nasal swab extract and processed for each study as described above. These dilutions were tested in triplicate. The lowest concentration demonstrating 3 of 3 positives was chosen for LoD confirmation. Based on this testing, the concentration chosen was of 0.88 x10 2 TCID
50 per mL.
LoD Confirmation
The analyte concentration 0.88 x10 2 TCID
50 per mL was tested twenty times to confirm. Twenty (20) of twenty (20) results were positive. Based on this testing the LoD was confirmed to be 0.88 x10 mL.
2 per
Cross-Reactivity
Cross-reactivity and potential interference of the Clip COVID Rapid Antigen Test was evaluated by testing 24 commensal and pathogenic microorganisms spiked into pooled human nasal wash, using the
Clip COVID Rapid Antigen Test. Each of the microorganisms were tested in triplicate in the absence or presence of inactivated SARS-CoV-2 at 3x LoD. No cross-reactivity or interference was seen with
Concentration Potential cross-reactant or interferent
Human Coronavirus 229E 1.52 x 10 5 TCID
50
/mL
Human Coronavirus OC43 5.05 x 10 4 TCID
50
/mL
Human Coronavirus NL63
Adenovirus Type 1
OKViruses
Human Metapneumovirus 9 (hMPV) Type A1
Parainfluenza Virus Type 1
Parainfluenza Virus Type 2
Parainfluenza Virus Type 3
1.71 x 10 4 TCID
50
/mL
1.03 x 10 7 TCID
50
/mL
1.18 x 10 4 TCID
50
/mL
3.42 x 10 6 TCID
50
/mL
5.05 x 10 4 TCID
50
/mL
8.58 x 10 6 TCID
50
/mL
Luminostics, Inc. | Page 13 of 16
Yeast
Bacteria
Parainfluenza Virus Type 4B
Influenza A H3N2 Brisbane/10/07
Influenza A H1N1 New Caledonia/20/99
Influenza B Brisbane/33/08
Enterovirus Type 68
Respiratory Syncytial Virus Type A (RSV-A)
1.16 x 10 6 TCID
50
/mL
3.55 x 10 4 TCID
50
/mL
4.17 x 10 4 TCID
50
/mL
1 x 10 4.07 TCID
50
/mL
1.51 x 10 5 TCID
50
/mL
4.17 x 10 4 TCID
50
/mL
1.6 x 10 7 TCID
50
/mL
5.43 x 10 7 CFU/mL
2.26 x 10 8 CFU/mL
Human Rhinovirus 17
Haemophilus influenzae
Type b Strain Egan
Streptococcus pneumoniae
Type 19F; Z022
Bordetella pertussis
Strain A639
Chlamydophila pneumoniae
Strain AR-39
Legionella pneumophila
Philadelphia
Mycoplasma pneumoniae
1.13 x 10 9 CFU/mL
1.4 x 10 7 IFU/mL
1.88 x 10 9 CFU/mL
Pneumocystis jiroveci
Recombinant W303-PJI 1.56 x 10 7 CFU/mL
Streptococcus pyogenes
2.66 x 10 8 CFU/mL
3.16 x 10 7 CCU/mL
Staphylococcus aureus 5.5 x 10 8 CFU/mL
Staphylococcus epidermidis
Candida albicans
7.7 x 10 8 CFU/mL
4.5 x 10 7 CFU/mL
Luminostics, Inc. | Page 14 of 16
Due to lack of availability for wet testing, the following pathogens were analyzed cross-reactive:
in silico
by comparing sequence homology on NCBI BLAST and determined to not be cross-reactive or highly unlikely to be
● MERS
● Coronavirus HKU1
●
M. tuberculosis
Due to lack of availability for wet testing, the following pathogens were analyzed determined to be cross-reactive:
in silico
and
• Severe Acute Respiratory Syndrome Coronavirus (SARS-CoV)
Endogenous Interference
The following 18 substances which can be expected to be naturally present in respiratory specimens or be artificially introduced, were evaluated with the Clip COVID Rapid Antigen Test at the concentrations listed below and were found not to affect test performance (i.e., they were found to not
Potential Interfering Substance Concentration in Sample
Zanamivir
Oseltamivir
282.0 ng/mL
2.2 μg/mL
Flonase
Saline nasal spray
0.7 g/mL
15% v/v
Rhinocort
Nasacort Allergy 24 hour
Afrin
Zicam Cold Remedy
Neo-Synephrine
Human Blood
Purified Mucin Protein
Tobramycin
Naso GEL (NeilMed)
5% v/v
5% v/v
5% v/v
5% v/v
5% v/v
5% v/v
5% v/v
2.5 mg/mL
1.25 mg/mL
Luminostics, Inc. | Page 15 of 16
CVS Nasal Spray (Cromolyn)
Homeopathic (Alkalol)
Sore Throat Phenol Spray (Chloraseptic)
Mupirocin
Fluticasone Propionate
15% v/v
10% v/v
15% v/v
10 mg/mL
5% v/v
High Dose Hook Effect
No high dose hook effect was observed when inactivated SARS-CoV-2 stock was tested at concentration of 9.58 x10 5 TCID50 per mL.
CLEANING AND DISINFECTING THE CLIP ANALYZER
Do not disassemble the Analyzer. The Analyzer contains no user-serviceable components.
Possible electrical shock: turn off and unplug the Analyzer prior to cleaning. Do not clean the port on the side of the instrument.
The Clip Analyzer can be gently wiped down with typical lab disinfectants (e.g., paper towel sprayed with 70% alcohol or Clorox/Lysol wipes) for cleaning if your protocols call for it. Do not spray disinfectant directly onto the Analyzer or immerse the Analyzer in liquid. Luminostics recommends disinfecting the Analyzer at least once per day.
ASSISTANCE
If you have any questions regarding the use of this product or if you want to report a test system problem, please contact Luminostics Technical Support at [email protected]
(phone: 800.FDA.1088; fax: 800.FDA.0178; http://www.fda.gov/medwatch).
. Test system
CONTACT INFORMATION
Luminostics, Inc.
446 South Hillview Drive
Milpitas, CA 95035 www.luminostics.com [email protected]
Luminostics, Inc. | Page 16 of 16
SYMBOLS
Manufacturer Catalogue
Number
For in vitro diagnostic use only
Batch Code
Do not reuse Consult instructions for use
Contains sufficient materials for 25 tests
Biohazard
© 2020 Luminostics, Inc. All rights reserved.
Prescription
Only
Use by date
Caution
Keep dry
Luminostics, Inc. | Page 17 of 16
Quick Reference
Instructions
For Use Under Emergency Use Authorization (EUA) Only. For in vitro Diagnostic Use. Rx Only.
Clip COVID Rapid Antigen Test Components use with the Clip Analyzer, is intended for protein antigen from SARS-CoV-2 in anterior nasal swab specimens from individuals who are suspected of COVIDbased on a diagnostics platform that uses nanoparticles, which are sensitively the qualitative detection of nucleocapsid
19 by their healthcare provider within the first five days of onset of symptoms.
Manufactured by Luminostics, Inc., it is proprietary persistent luminescent detectable using a smartphone’s camera, and the Clip COVID App. as lateral flow assay reporters. The Clip
Analyzer is comprised of an iPhone SE, an
Adapter (pre-assembled onto the iPhone),
Extraction
Tube
Clip Analyzer
Dropper Tip
Cartridge
Nasal Swab
Study the Clip Analyzer User Manual and Clip COVID Rapid Antigen Test Kit Package Insert thoroughly before using these Quick Reference Instructions or performing a test. This is not a complete product insert. Perform the Clip COVID Rapid Antigen Test at room temperature between
15 ° C and 30 ° C (59 ° F and 86 ° F). Nasal swab specimens must be processed within 30 minutes of collection. Specimens and kit components must be at room temperature before testing. Check expiration date on outer test kit carton and each individual test package before using. Do not use any test component beyond its expiration date. Refer to the Clip COVID Rapid Antigen Test Kit
Package Insert for Specimen Collection, Warning and Precautions, and Limitations.
TEST PROCEDURE
1
Place the Clip Analyzer on a table or countertop and power it on by holding down the power button on the right side of the iPhone. The Analyzer is portable and can be moved to a suitable location for testing. Ensure the surface is stable, level, dry and free of obstructions.
Ensure the surface provides adequate space for the Clip Analyzer. There must be space to access the Clip Analyzer port for insertion of the Cartridge. We recommend that you keep the Analyzer plugged in to a power outlet using the provided charging cord during operation/testing.
2 Remove a Cartridge from its foil pouch after using the tear notch to open the pouch.
3
Load the Cartridge into the Analyzer by pushing the Cartridge into the
Cartridge port until you hear a click. If you don’t hear a click, continue pushing the Cartridge until you can’t push it further.
4
Touch ‘Begin Test’ on the home screen of the Clip
COVID app on the
ID and select the Test Kit
Analyzer. Enter sample lot number — printed on the Cartridge foil pouch label — from the dropdown menu. Touch
‘Continue’.
Touch “Begin Test”
In the USA, this test has not been FDA cleared or approved, but has been authorized by FDA under an EUA for emergency use by laboratories certified under the CLIA, 42 U.S.C. moderate, or waived complexity tests. This test is for use at the Point of Care (POC), i.e., in patient care settings operating under a CLIA Certificate of Waiver, Certificate of Compliance, or Certificate of Accreditation. This test has justifying the authorization of emergency use of in vitro diagnostics for detection and/or diagnosis of the virus that causes COVID-19 under Section 564(b)(1) of the Federal Food, Drug and Cosmetic Act, 21 U.S.C.
USA, the emergency use of this test is only authorized for the duration of the declaration that circumstances exist
§ 263a, that meet the requirements to perform high, unless the declaration is terminated or authorization is revoked sooner.
§ 360bbb-3(b)(1),
5
Confirm the adapter ID by the ID on screen does not match the ID on the side of adapter to the unit ID on the screen. Touch ‘Confirm’ if the ID on screen matches the ID visible on the
Analyzer. Touch ‘Retry’ if the Analyzer. a few seconds
6
Insert the nasal swab from the patient all the way into the bottom Extraction Tube
Squeeze the center of the tube. Leave the swab in the buffer in the Extraction
Tube for 60 seconds.
Extraction Tube and remove the swab while keeping the center of the tube squeezed. Dispose of swab in a biohazard waste stream. Cap Extraction
Tube using the Dropper Tip.
7
Dispense the entirety of the contents of the Extraction
Tube into the sample well of the Cartridge by turning it upside down and squeezing it. Holding the tube vertically directly above the sample well will minimize spillage.
X
3
+
8
The Analyzer will
“invalid” result will display in disturb the Analyzer until a
30-45 seconds after sample addition, transitioning to the
‘Analysis in Progress’ screen.
30 minutes. Do not touch or remove the Cartridge or
RESULT INTERPRETATION
Positive. SARS-CoV-2 antigen present; does not rule out coinfection with other pathogens.
Negative. Negative results from patients with symptom onset beyond five days should be treated as presumptive and confirmation with a molecular assay, if necessary, may be performed. treatment or patient management decisions, including infection control decisions.
Negative results do not rule out SARS-CoV-2 infection and should not be used as the sole basis for
Invalid. Should this occur, retest patient with a new nasal swab, Cartridge, and Extraction Tube.
QUALITY CONTROLS
It is recommended that positive and negative external control swabs are run once by each untrained operator and as deemed additionally necessary by your internal quality control procedures and in accordance with Local, State and Federal regulations or accreditation
Card or Package Insert. Patient tests should not be performed if the test fails to detect positive and/or negative control swabs accurately; please re-test or contact Luminostics, Inc. requirements. External positive and negative control swabs are provided in the kit. The external controls should be tested using the test procedure provided in this Quick Reference Instructions
CUSTOMER SERVICE
If the Clip COVID Rapid Antigen Test Kit or Clip Analyzer do not perform as expected, contact
Luminostics, Inc. at [email protected].
MANUFACTURER INFORMATION
Designed & manufactured by Luminostics, Inc. at 446 South Hillview Drive, Milpitas, CA 95035, USA.
Copyright © 2020 Luminostics, Inc. v1.2 revised 12-04-20.
User Manual — Clip Analyzer
For Use With:
Clip Analyzer User Manual
For Use With Clip COVID Rapid Antigen Test
For in vitro Use Only
Rx Only
For Use Under Emergency Use Authorization (EUA) Only
INTENDED USE
The Clip Analyzer is an analyzer intended to be used for objective readout of Cartridge-based immunoluminescent
in vitro
diagnostic assays manufactured by Luminostics. The Clip Analyzer is intended for professional and laboratory use. The first in vitro diagnostic test made available for use on the Clip Analyzer is the Clip COVID Rapid Antigen Test under FDA emergency use authorization
(EUA).
PRODUCT DESCRIPTION
The Clip Analyzer comprises an Apple iPhone SE(2020), an Adapter (pre-assembled onto the iPhone), and the Clip COVID iOS app. The iPhone has been delivered to you in single-app mode, i.e., it is only capable of running the Clip COVID app.
SYSTEM COMPONENTS
The following system components are supplied with the Clip Analyzer:
● Analyzer
● Charging Cord and Power Adapter
● User Manual separately. Contact Luminostics Technical Support for additional supplies at [email protected].
WARNINGS AND PRECAUTIONS
● For
●
in vitro
diagnostic use.
For prescription use only
● use by FDA under an Emergency Use Authorization (EUA) for use by laboratories certified
Luminostics, Inc.
| Page 1 of 13
Clip Analyzer User Manual
For Use With Clip COVID Rapid Antigen Test under the CLIA, 42 U.S.C. §263a, that meet the requirements to perform high, moderate or waived complexity tests.
● This product has been authorized only for the detection of proteins from SARS-CoV-2, not for any other viruses or pathogens.
● The emergency use of this product is only authorized for the duration of the declaration that circumstances exist justifying the authorization of emergency use of in vitro diagnostic tests for detection and/or diagnosis of COVID-19 under Section 564(b)(1) of the Federal Food, Drug, and Cosmetic Act, 21 U.S.C. § 360bbb-3(b)(1), unless the declaration is terminated or authorization is revoked sooner.
● Use of Nitrile, Latex (or equivalent) gloves is recommended when handling patient samples.
Change gloves between patients.
● Always operate the Clip Analyzer and use other components of the Clip COVID Rapid
Antigen Test on a surface that is level, dry, and not in direct sunlight. Failure to do so may sample processing. cause false positive, false negative, or invalid results. Leave room around the Analyzer for
● Do not move or adjust the Clip Analyzer or remove the Cartridge while there is a test in progress.
Doing so may cause an invalid result.
● Use only the power adapter that was provided with the Analyzer.
● Do not drop the Analyzer, as it could damage the unit.
● To avoid damaging the Analyzer, do not place objects on top of it.
● Dispose of containers and unused contents in accordance with Federal, State, and Local regulatory requirements. The used nasal swab, cartridge, extraction tube, and dropper tip are considered biohazardous waste and should be disposed of in a manner consistent with local biohazard waste disposal regulations.
● Do not spray disinfectant directly onto the analyzer.
● The product has not been tested for EMI compatibility with implantable cardioverterdefibrillators (ICDs) or pacemakers. Do not use the Clip Analyzer if you have an ICD or pacemaker.
● Do not open the foil pouch of the Cartridge and expose it to the ambient environment until the Cartridge is ready for immediate use. Premature exposure to ambient conditions may cause false positive, false negative, or invalid results.
SYSTEM INSTALLATION, SETUP, AND OPERATION
Analyzer Setup to a suitable location for testing, ideally near an electrical outlet for charging. Ensure the counter top is stable, level, dry and free of obstructions. Avoid direct sunlight. Ensure the bench provides adequate space for Clip Analyzer. There must be space to insert the Cartridge into the Analyzer. Plug the
Charging Cord into the iPhone’s charging port on the bottom of the Analyzer. Then plug the Charging
Power Up
Luminostics, Inc.
| Page 2 of 13
Turn on the iPhone sub-component of the Analyzer by depressing the button on the right side of the bezel. Upon insertion of the Cartridge, the Analyzer will turn on.
Run Test
Clip Analyzer User Manual
For Use With Clip COVID Rapid Antigen Test
Follow the assay-specific Package Insert to run a test using the Clip Analyzer.
Patient Test Result
When the test is complete, the results for the patient specimen will be displayed on the Analyzer screen.
Shutdown
Turn off the unit by removing the cartridge and holding the power switch on the right side of the unit.
Shutdown is complete when the screen goes dark.
STORAGE AND OPERATING CONDITIONS
Store and operate the Clip Analyzer at room temperature, 59°F to 86°F (15°C to 30°C), out of direct sunlight, between 20%-85% humidity (non-condensing).
TEST PROCEDURE FOR CLIP COVID RAPID ANTIGEN TEST
1. Place the Clip Analyzer on a table or counter top and power it on as instructed in the testing. Ensure the surface is stable, level, dry and free of obstructions. Ensure the bench plugged in to a power outlet using the provided charging cord during operation/testing.
Analyzer manual. The Analyzer is portable and can be moved to a suitable location for provides adequate space for the Clip Analyzer. There must be space to access the Clip
Analyzer port for insertion of the Cartridge. We recommend that you keep the Analyzer
2. Remove a Cartridge from its foil pouch after using the tear notch to open the pouch.
3. Load the Cartridge into the Analyzer by pushing the Cartridge into the Cartridge port until you hear a click. If you don’t hear a click, continue pushing the Cartridge until you can’t push
Luminostics, Inc.
| Page 3 of 13
Clip Analyzer User Manual
For Use With Clip COVID Rapid Antigen Test it further. The cartridge can be handled with bare hands. However, we recommend wearing gloves during the entirety of the test procedure including in this step.
4. Touch ‘Begin Test’ on the home screen of the Clip COVID app on the Analyzer
‘Continue’.
5. Enter sample ID and select the Test Kit lot number from the drop-down menu. Touch
6. Confirm the adapter ID by matching the label on the side of the adapter to the unit ID on the screen. Touch ‘Confirm’ if the ID on screen matches the ID on the side of the Analyzer.
7. Insert the anterior nasal swab collected from the patient all the way into the bottom
Extraction Tube and rotate the swab at least 3 times against the bottom of the tube.
8. Leave the swab in the buffer in the Extraction Tube for 60 seconds.
Luminostics, Inc.
| Page 4 of 13
Clip Analyzer User Manual
For Use With Clip COVID Rapid Antigen Test
9. Squeeze the center of the Extraction Tube and remove the swab while keeping the center of the tube squeezed. Dispose of swab in a biohazard waste stream.
11. Dispense the entirety of the contents of the Extraction Tube into the sample well of the
Cartridge by turning it upside down and squeezing it. Holding the tube vertically directly above the sample port will minimize spillage.
12. The Analyzer will automatically begin analysis 30-45 seconds after sample addition, transitioning to the ‘Analysis in Progress’ screen. A “positive”, “negative”, or “invalid” result a result is displayed. Movement of or removal the cartridge while analysis is in progress will will display in 30 minutes. Do not touch or remove the cartridge or disturb the Analyzer until result in invalid result.
Luminostics, Inc.
| Page 5 of 13
Clip Analyzer User Manual
For Use With Clip COVID Rapid Antigen Test
RESULT INTERPRETATION
When the test is complete, the result will be displayed on the Analyzer screen. The result of the lateral flow test cannot be seen with the naked eye. The Analyzer screen will display results, individually providing a positive or negative result for SARS-CoV-2. If the result is Invalid, retest with a new patient
Positive Result Negative Result Invalid Result
This display shows a valid positive result for SARS-CoV-2.
This display shows a valid negative result for SARS-CoV-
2.
This display shows an invalid result.
LIMITATIONS
● This product is currently authorized to only be used for the qualitative detection of SARS-
●
CoV-2 antigens from an anterior nasal swab.
The Clip COVID Rapid Antigen Test detects both viable (live) and non-viable SARS-CoV-2.
Test performance depends on the amount of virus (antigen) in the sample and may or may not correlate with viral culture results performed on the same sample.
● A negative test result may occur if the level of antigen in a sample is below the detection limit of the test or if the sample was collected improperly.
● Failure to follow the Test Procedure may adversely affect test performance and/or invalidate the test result by causing false positive or false negative results .
● Test results must be evaluated in conjunction with other clinical data available to the physician.
● Positive test results do not rule out co-infections with other pathogens.
● Positive test results do not differentiate between SARS-CoV and SARS-CoV-2.
● Negative test results are not intended to rule in other non-SARS viral or bacterial infections.
Luminostics, Inc.
| Page 6 of 13
Clip Analyzer User Manual
For Use With Clip COVID Rapid Antigen Test
● Negative results, from patients with symptom onset beyond five days, should be treated as
● If the differentiation of specific SARS viruses and strains is needed, additional testing, in
● performed for clinical management. consultation with state or local public health departments, is required.
Modifications to the assay procedure may result in false positive or false negative results. presumptive and confirmation with an FDA authorized molecular assay, if necessary, may be
● False negative results may occur if a specimen is improperly collected, transported or handled.
● The test has not been validated for use with viral transport media (VTM) or universal transport media (UTM). Usage of the test with samples prepared using VTM or UTM may cause false positive, false negative, or invalid results.
● The test has been validated for use in temperatures ranging from 15 ° C-30 ° C. The test has not been validated for use in temperature ranges outside of these conditions and usage outside of
CONDITIONS OF AUTHORIZATION FOR THE LABORATORY
The Clip COVID Rapid Antigen Test Letter of Authorization
, along with the authorized Fact Sheet for Healthcare Providers, the authorized Fact Sheet for Patients, and authorized labeling are available on the FDA website: https://www.fda.gov/medical-devices/emergency-situations-medical-devices/emergency-useauthorizations#covid19ivd.
However, to assist clinical laboratories using the Clip COVID Rapid Antigen Test (“your product” in the conditions below), the relevant Conditions of Authorization are listed below:
● Authorized laboratories using your product will include with test result reports, all authorized
Fact Sheets. Under exigent circumstances, other appropriate methods for disseminating these
Fact Sheets may be used, which may include mass media.
● Authorized laboratories using your product will use your product as outlined in the “Clip
COVID rapid antigen” Package Insert and User Manual. Deviations from the authorized procedures, including the authorized instruments, authorized clinical specimen types, authorized control materials, authorized other ancillary reagents and authorized materials required to use your product are not permitted.
● Authorized laboratories that receive your product will notify the relevant public health
● authorities of their intent to run your product prior to initiating testing.
Authorized laboratories using your product will have a process in place for reporting test results
1 The letter of authorization refers to, “Laboratories certified under the Clinical Laboratory
Improvement Amendments of 1988 (CLIA), 42 U.S.C. §263a, that meet the requirements to perform high, moderate, or waived complexity tests. This test is authorized for use at the Point of Care (POC), i.e., in patient care settings operating under a CLIA Certificate of Waiver, Certificate of Compliance, or Certificate of Accreditation.” as “authorized laboratories.”
Luminostics, Inc.
| Page 7 of 13
Clip Analyzer User Manual
For Use With Clip COVID Rapid Antigen Test
● Authorized laboratories will collect information on the performance of your product and report to DMD/OHT7-OIR/OPEQ/CDRH (via email: [email protected]) and
Luminostics (via email: [email protected]), any suspected occurrence of false positive or false negative results and significant deviations from the established performance characteristics of your product of which they become aware.
● All operators using your product must be appropriately trained in performing and interpreting the results of your product, use appropriate personal protective equipment when handling this kit, and use your product in accordance with the authorized labeling.
● Luminostics, Inc., authorized distributors, and authorized laboratories and patient care settings using your product will ensure that any records associated with this EUA are maintained until otherwise notified by FDA. Such records will be made available to FDA for inspection upon request.
BATTERY POWER
Battery Intended Use
Clip Analyzer includes 2 disposable AA batteries to power the Adapter and the iPhone SE(2020)’s rechargeable battery that is recharged when connected to AC power.
Battery Replacement
The iPhone sub-component of the Clip Analyzer is shipped with an internal LiPo rechargeable battery with an expected life of approximately three years. 2 disposable AA batteries with an expected life of
450 tests are used to power the non-iPhone sub-component of the Analyzer.
The disposable AA batteries are user-replaceable. The internal rechargeable battery is not user replaceable.
Prior to replacing the disposable AA batteries, ensure that there is no cartridge in the Analyzer. Remove the battery cover on the rear of the Analyzer. Carefully remove the used batteries and insert the new ones.
Replace the battery cover. Recycle or dispose of the batteries in accordance with all Federal, State and
Local laws. To avoid fire and explosion hazard, do not burn or incinerate the batteries.
MAINTENANCE
The Clip Analyzer must be sent to Luminostics if maintenance is required. The user should not attempt any maintenance except for changing the batteries and cleaning the external surfaces. Contact
Luminostics Technical Support via email at [email protected] for maintenance, return, or disposal of Clip Analyzer.
Luminostics, Inc.
| Page 8 of 13
Clip Analyzer User Manual
For Use With Clip COVID Rapid Antigen Test
CLEANING AND DISINFECTING THE CLIP ANALYZER
Do not disassemble the Analyzer. The Analyzer contains no user-serviceable components.
Possible electrical shock: turn off and unplug the Analyzer prior to cleaning. Do not clean the port on the side of the instrument.
The Clip Analyzer can be gently wiped down with typical lab disinfectants (e.g., paper towel sprayed with 70% alcohol or Clorox/Lysol wipes) for cleaning if your protocols call for it. Do not recommends disinfecting the Analyzer at least once per day.
POTENTIAL BIOHAZARD
Dispose of used specimens in accordance with Federal, State, and Local requirements for biohazard waste.
Treat specimens and patient samples as potentially biohazardous material.
Ensure the Analyzer is cleaned per the Cleaning Guidelines.
Seek specific training or guidance if you are not experienced with specimen collection and handling procedures.
Please read the entirety of this user manual and the package insert prior to executing a test.
Use of Nitrile, Latex, or other gloves is recommended when handling patient samples. Change gloves between patients.
ASSISTANCE
If you have any questions regarding the use of this product or if you want to report a test system problem, please contact Luminostics Technical Support at [email protected]
(phone: 800.FDA.1088; fax: 800.FDA.0178; http://www.fda.gov/medwatch).
. Test system
CONTACT INFORMATION
Luminostics, Inc.
446 South Hillview Drive
Milpitas, CA 95035 www.luminostics.com [email protected]
Luminostics, Inc.
| Page 9 of 13
SYMBOLS
Manufacturer
For in vitro diagnostic use only
Keep dry
MENU STRUCTURE
Catalogue
Number
Clip Analyzer User Manual
For Use With Clip COVID Rapid Antigen Test
Prescription
Only
Serial
Number
Biohazard
Testing Menu
Consult instructions for use
Caution
Luminostics, Inc.
| Page 10 of 13
Clip Analyzer User Manual
For Use With Clip COVID Rapid Antigen Test
Cancellation Pop-up
TROUBLESHOOTING
Smartphone Low
Battery
This screen is shown when the smartphone does not have sufficient power.
Please plug in the charging cable.
Adapter Low Battery
This screen is shown when the adapter has a low battery, approximately 20%.
You may be able to complete a test.
Changing batteries is recommended.
Luminostics, Inc.
| Page 11 of 13
Adapter Critical
Battery
This screen is shown when adapter battery levels are critical, at approximately 5%.
You will not be able to complete a test.
Batteries should be replaced.
Adapter Not Found
The Adapter is not sensed by the smartphone. Please check the Adapter batteries, confirming that the Adapter has power and a cartridge is inserted.
Clip Analyzer User Manual
For Use With Clip COVID Rapid Antigen Test
Cartridge Removed
The cartridge has been removed from the
Analyzer. The current test run and cartridge are invalid.
To execute a test, take another sample and, execute a test.
Motion Detected
The Analyzer has sensed motion. Please secure the device in a stable location. The test should automatically resume.
Adapter Disconnected
The Adapter is not sensed by the smartphone. Please check the Adapter batteries, confirming that the Adapter has power and a cartridge is inserted.
The current test run and cartridge are invalid.
Luminostics, Inc.
| Page 12 of 13
Clip Analyzer User Manual
For Use With Clip COVID Rapid Antigen Test
Take another sample and, using a new cartridge, execute a test.
TECHNICAL SPECIFICATIONS
Power Supply 1.9-3.5V DC, Max.
0.14A from 2 AA batteries.
Dimensions
Weight
77x37x155mm
336g
Display
Operational Temperature
Operational Humidity iPhone SE (2020), 5.45 x 2.65 in
15°C to 30°C
20%-85% non-condensing
The Clip Analyzer will be tested to UL 60601-1 3rd Edition.
© 2020 Luminostics, Inc. All rights reserved.
Luminostics, Inc.
| Page 13 of 13