BMC RESmart is a multifunctional device for treating obstructive sleep apnea. It utilizes continuous positive airway pressure (CPAP) therapy to maintain open airways during sleep. It features an adjustable pressure range, humidifier, and user-friendly interface. Its advanced algorithms monitor and respond to individual breathing patterns, providing personalized therapy.
Reklama
Reklama
RESmart
®
CPAP and AutoCPAP System
Service Manual
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
Table of Contents
1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2. Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.1 Outer Illumination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2 Inner Construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2 Spare Part . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3. Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.1 Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.2 Hidden Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4. Malfunction and Countermeasure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4.1 Malfunction Determination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4.2 Error List and Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Appendices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Appendix A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Appendix B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
1. Introduction
The RESmart ® CPAP / Auto-CPAP units are continuous positive airway pressure devices designed for the treatment of adult obstructive sleep apnea (OSA) only.
This service manual is intended only for manufacturer authorized service personnel. Maintenance actions performed by unauthorized service personnel will serve to void the manufacturer’s warranty.
3
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
2. Structure
2.1 Outer Illumination
Handle
DisplayScreen
User Buttons
Humidifier
Controller
Air Outlet
Humidifier
Power
Outlet
Port Indicator
Light
Heater
Plate
Fill Line
Water
Chamber
Medical
Product Note
(bottom)
AC Inlet
SD Card
Filter Cap
& Filter
2.2 Inner Construction
See appendix A
2.3 Spare Part
See appendix B
4
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
3. Software
3.1 Instruction for Upgrading Device Software
1. Request copy of upgrade file prior to starting your upgrade, copy on to new
SD card.
2. Power off RESmart® CPAP/AutoCPAP device, then insert new SD Card.
3. Press and hold both Pressure Start/Stop Button and User Button (-). Plug in power cord while keeping both buttons pressed.
4. System will begin upgrade.
5
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
3.2 Hidden Function
1. Clearing Patient Data
(Available on version 1.28 and later)
Clearing the data of the RESmart CPAP and Auto-CPAP unit requires use of a hidden function. Caution: Once the data is cleared, there is no possibility of recovering erased data.
To clear all recorded data and restore the unit to initial factory default settings, enter the Patient Menu by depressing ramp for 3 seconds or more. Advance the menu using the “+” sign until you reach the date option. Manually change the date to reflect: Year = 2097, Month = 07, Day =01, Hour = 0, Minute =26.
Press On/Off button when “Hour” setting is displayed. It is not necessary to save the setting.
Important note: After data has been cleared, it will be necessary to
re-set the device, reviewing all patient and clinician settings (i.e. enabling iCode, correcting to current date, etc.).
6
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
2. Pressure calibration
(Available on version 1.20 and later)
Connect pressure meter to the gas outlet in the device, enter hidden datesetting, set ‘YYYY/MM/DD’=’2098/08/08’, press ‘On/Off’ when ‘Hour’ setting, thus pressure calibration starts. The device increases pressure to
20hPa(cmH2O), AD sampling data will appear on LCD. When LCD data displaying is stable, if pressure meter value is not 20, press ‘+/-’ button to adjust the pressure output. Until output pressure value is 20 and LCD displays stable, press ‘On/Off’ button to finish pressure calibration.
3. Debug mode
(Available on version 1.19 and later)
Enter hidden date setting, set ‘YYYY/MM/DD’=’2099/08/02(or 03)’, press ‘On/
Off’ when ‘Hour’ setting, thus debug #02 or #03 mode starts. The device can send real time data to computer via serial port. Only cutting off power supply can make device back to normal mode.
7
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
4. Malfunction and Countermeasure
4.1 Malfunction Determination
8
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
4.2 Error List and Information
Error
Index Problem
01 Overheat
Causing
Error reading from temperature sensor.
Possible temperature sensor damage. Reset device if singular occurrence.
20
21
22
23
24
64
74
75
09
12
13
14
15
16
17
18
19
02 Blower feedback error
03 Blower runs slowly
04 Temperature sensor is damaged.
07
08
Zero point drifting of pressure sensor ultra limit
Version mismatch
Device parameter error
Sensor error
Device temperature is lower than 0℃
Lack of hardware parameter
BPAP software error
BPAP software error
Blower driver circuit overheats.
Check blower and pressure sensor.
Blower, main board or pressure sensor problem.
Parameter error or pressure sensor damaged.
The system was updated by a wrong version.
Check device driver and circuit board
The device’s temperature is too high or digital pressure sensor’s problem.
Digital pressure sensor problem.
Blower overheats The temperature of blower is too high or the blower temperature sensor is damaged.
Possible damage to temperature sensor Restart device if singular occurrence.
Unexpected error generated from sensor
Write the software of 25T/30T into 25/25A.
Write the software of 25/25A into 25T/30T.
The temperature sensor of blower driver circuit is damaged.
Two sensors overheat at the same time The main board and flow sensor overheat
(BPAP, Auto
Blower current is over flow.
The mean current of blower is over flow.
at the same time. CPAP: The main board and the thermostat of power tube overheat at the same time.)
During pressure calibration, the mean current is over flow.
3.3V voltage supplied to digital flow sensor is abnormal.
Blower data from serial port is wrong during calibration.
Blower cannot run successfully in 20 seconds
9
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
Appendices
Appendix A:
Device Configuration
1. Main device
Fig. A - Main board (backside) Fig. B – Outlet (leftside)
Hole #1, 2 and 3 in Fig. B, connect to sensor YL and LL via silicon rubber canal.
Specified in Table 1:
2
3
Fig. B
1
Fig. A
Hole #1 on sensor YL
Hole #2 on sensor LL
Hole #1 on sensor LL
Table 1
Specification of canal
℃2mm, ℃4mm, L12cm
℃2mm, ℃4mm, L12cm
℃2mm, ℃4mm, L8cm
10
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
Fig. C – Power PCBA (front side)
Sockets figuration
Fig. D – Wire connecting on
power PCBA
AC Socket
Blue
Brown
3-pin plug
6-pin plug
Orange
Orange
Black
Black
Fig. E – Power supply connecting
2-pin plug Reed
11
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
Fig. F – Power supply to humidifier
Connecting relationship:
Table 2
12
2. Humidifier
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
Fig. G – Humidifier PCBA (back side)
Fig. H – Power supply to humidifier
13
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
Fig. I – Heater PCBA
4-pin plug
3-pin plug
Power Reed to Humidifer
Heater PCBA
Connecting relationship:
Table 3
14
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
Components Illumination
1. Device Inside
S/N Assembly
1 Insert humidifier locker on the shield bottom.
Figure
Orange Black
2 Insert humidifier power supply reed on the shield bottom as per Fig. E and F.
3 Assembly the humidifier locker and button spring.
4 Put fixing on the spring and fix by screw (2×3mm*8mm).
5 Put power wire.
6 Assembly the shield back on the bottom, fix by screw
(2×3mm*8mm).
7 Assembly foam (bottom) into shield bottom.
8 Assembly AC power socket on the shield back (character face up).
Power wire to humidifier
15
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
S/N Assembly Figure
9 Clean up the wire.
10 Assembly power PCBA in the socket and connect all plugs. Clean up the wire.
11 Connect outlet to blower connector, fix by clip.
12 Connect blower to connector and fix by clip. Put the integrated one in the right place.
13 Covered by foam cover and clean up wires.
14 Assembly fixing clip on the main board PCBA.
15 Plug the 8-pin plug of blower in the socket PM plug #2 as in
Fig. E in the socket P24 on the main board PCBA.
FCBA Fixing
Clip
Pressure Sensor
16
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
S/N Assembly Figure
FCBA Fixing Pole
16 Assembly the main board PCBA on fixing pole and make sure the clip clamps power PCBA.
17 1. As per Table 1, connect sensor
YL to outlet by silicon rubber
canal.
2. Connect sensor LL to outlet by
silicon rubber canal.
12cm canal for YL Sensor
8cm canal for LL Sensor
12cm canal for LL Sensor
2. Device Outside
S/N Assembly
18 Put handle in the socket on shield cover.
Figure
17
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
S/N Assembly Figure
19 Assembly handle fixing to handle and fix by screw (4×3mm*8mm).
20 Assembly the infrared window.
21 Assembly and fix the panel.
Infrared
Window
23 Put on the key button.
24 Assembly shield cover on the bottom, adjust the position of outlet and make sure it is clamped in the right socket.
25 Fix the shield by screws
(4×M3*10mm).
26 Stick four device pads.
27 Put in filter and assembly the cover.
Screw Hole
Device Pad
18
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
S/N Assembly Figure
28 Put on all plugs if necessary
(without humidifier).
Plug (up)
Power Plug
Plug (down)
2. Humidifier
S/N Assembly
1 As per Fig. H, assembly the heater plane and wire. Fix them by screw (3mm*8mm).
Black
Figure
Orange
2 Put spring on each of three poles on heater plane.
3 Assembly the scaleboard on the heater plane and fix by screw
(3mm*8mm).
4 As per Table 3, assembly heater
PCBA and wires.
5 Put the humidifier inner connector on t he shield.
6 Put on light window.
Scaleboard
19
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
Figure S/N Assembly
7 Assembly the inner and outer
shield.
8 Fix on screw (M3*10mm).
Screw
20
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
Appendix B:
Spare Part Configuration
S/N Name Qty Figure
1 Shield Cover 1
2 Handle
3 Handle Fixing
4 Shield Bottom
1
2
1
5 Humidifier Clip 1
6 Humidifier Locker 1
7 Humidifier Fixing 1
8 Filter Cover
9 Shield Back
1
1
21
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
S/N
10 Outlet
Name Qty Figure
1
11 PCBA Fixing
12 Infrared Window
13 Panel
14 Humidifier Outer Shield
15 Humidifier Inner Shield
16 Light Window
17 Pole Platelet
18 Humidifier Power Plug
19 Humidifier Plug (left-up)
20 Humidifier Plug (left-down) 1
1
1
1
1
2
2
1
1
1
22
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
S/N Name
21 Humidifier Plug (right-up)
Qty Figure
1
22 Humidifier Plug (right-down) 1
23 Device Pad
24 Blower Connector
6
1
25 Humidifier Inner Connector 1
26 Key Button
27 Humidifier Scaleboard
28 Heater Plane Spring
29 Button Spring
1
1
3
1
23
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
Spare Part List
Part Name S/N
Main Board PCBA
Heater PCBA
Power Supply PCBA
Blower
Heater Plane
Power Supply PCBA Wire
Main Board PCBA Wire
Silicon Rubber Canal
Silicon Rubber Canal
Foam (bottom)
Foam (cover)
Shield Cover
Handle
Handle Fixing
Shield Bottom
Humidifier Clip
Humidifier Locker
Humidifier Fixing
Filter Cover
Shield Back
Outlet
PCBA Fixing
Infrared Window
Panel
40111-0601
40111-0701
4011120801
4011120901
4011123001
40111-1301
4011101401
4011220101
Power Supply PCBA Fixing 40111-1601
Humidifier Outer Shield 4011122201
Humidifier Inner Shield
Pole Platelet
4011122301
40111-2601
Light Window
Button Spring
Humidifier Scaleboard
Heater Plane Spring
4011142401
4013140101
4013130401
4013140501
20211261
20212-41
20311101
40211
45010101
45010201
45010301
50224
50224
40191-0101
40191-0201
4011120101
4011120201
40111-0301
4011120401
4011120501
Humidifier Power Plug
Humidifier Plug (left-up)
4012110101
4012110201
Humidifier Plug (left-down) 4012110301
Humidifier Plug (right-up) 4012110401
Humidifier Plug (right-down) 4012110501
Device Pad 4012110601
Key Button
Blower Connector
4012110701
4012110801
Qty
1
1
1
3
1
1
1
1
1
2
2
1
1
1
1
1
1
6
1
1
1
1
1
1
1
2
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Unit Classify Use
PC
PC
PC
PC
PC
PC
PC
PC
PC
PC
PC
PC
Set
PC
PC
PC
PC
PC
PC
PC
PC
PC
PC
PC
PC
PC
PCBA
PCBA
Main Device
Humidifier
PC PCBA Main Device
PC Assembly Main Device
Set Assembly Humidifier
Set Wire Main Device
Set Wire Main Device
PC Assembly Main Device
PC Assembly Main Device
PC Foam Main Device
PC
PC
Foam
Plastic
Main Device
Main Device
PC
PC
PC
PC
Plastic
Plastic
Plastic
Plastic
Main Device
Main Device
Main Device
Main Device
Plastic Main Device
Plastic Main Device
Plastic Main Device
Plastic Main Device
Plastic Main Device
Plastic Main Device
Plastic Main Device
Plastic Main Device
Plastic Main Device
Plastic Humidifier
Plastic
Plastic
Plastic
Metal
Metal
Metal
Rubber
Rubber
Rubber
Rubber
Rubber
Rubber
Humidifier
Humidifier
Humidifier
Main Device
Humidifier
Humidifier
Main Device
Main Device
Main Device
Main Device
Main Device
All
Rubber Main Device
Rubber Main Device
24
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
Spare Part List
(cont.)
Part Name S/N
Humidifier Inner Connector 4012110901
Screw-1
Screw-2
Sunk Screw
Filter
Power Cord 33011101
Qty Unit Classify Use
1 PC Rubber
8 PC Screw
Humidifier
All
9 PC Screw
4 PC Screw
All
Main Device
1 PC Accessory
1 PC Accessory
All
All
25
Service Manual for RESmart ® CPAP and AutoCPAP V1.5
Limited Warranty
3B Medical, Inc. (hereafter ‘3B’) warrants that the RESmart ® CPAP and Auto-CPAP devices will be free of all defects in workmanship and materials, and will perform according to specifications, for a period of two (2) years from the date of sale of the device, and 90 days on any accessories.
If the product fails to perform in accordance with the product specifications,
3B will repair or replace at its option, the defective material or part. This warranty does not cover damage caused by accident, misuse, abuse, alteration and other defects not related to material or workmanship.
• 3B will issue an RMA (Return Merchandise Authorization) within 24 hours of receipt of written notification of a failed or defective unit.
Failed/defective units must be returned within 30 days of the RMA date.
• Shipping the machine to us for warranty diagnostic is the customer’s responsibility. Warranty covers Domestic Shipping back to you.
Please contact us for all warranty claims or questions.
• This warranty coverage is applicable to all 3B/BMC CPAP, Auto
CPAP and Auto Bi-level devices.
• The warranty policy does not cover any damages caused as a result of alteration, intentional damage, modification, or unauthorized repair of the device.
• Warranty does include reasonable water damage and is at the sole discretion of 3B.
• This warranty replaces all other expressed or implied warranties, including any implied warranty of merchantability or fitness for a particular purpose. 3B reserves the right to amend this policy at any time.
3B Medical, Inc.
21301 Highway 27 N.
Lake Wales, FL 33859 USA
Email: [email protected]
Tel: (863) 226-6285 • Fax: (863) 226-6284
For additional information, please visit our Patient Portal at: www.3Bproducts.com