advertisement
530 S/SN Operating Manual
Contents
2 Declaration of incorporation
4 Information for returning pumps
5 Peristaltic pumps - an overview
9 Good pump installation practice
10.1 Keypad Layout and Key IDs
10.5 Change rotation direction
11 Connecting to a power supply
13 Switching the pump on for the first time
13.1 Selecting the display language
13.2 First-time start-up defaults
m-530sn-gb-02
1
2
14 Switching the pump on in subsequent power cycles
16.3 INCREASING AND DECREASING FLOW RATE
16.4 MAX FUNCTION (Manual mode only)
17.1 Setting the flow calibration
25.2 Tubing and element part numbers
m-530sn-gb-02
m-530sn-gb-02 3
4
1 Declaration of conformity
Watson-Marlow Limited
Falmouth
Cornwall
TR11 4RU
England
1. 530 Cased pumps (Models: S, SN, U, UN, Du, DuN, Bp, BpN)
2. Manufacturer:
Watson Marlow Ltd
Bickland Water Road
Falmouth
EC Declaration of Conformity
TR11 4RU
UK
3. This declaration of conformity is issued under the sole responsibility of the manufacturer
4. All models and versions of the 530 series of cased peristaltic pump with all approved pump heads, tubing and accessories.
5. The object of the declaration described above is in conformity with the relevant Union harmonisation legislation:
Machinery Directive 2006/42/EC
EMC Directive 2014/30/EC (from 20 th
ROHS Directive 2011/65/EU
April 2016) and 2004/108/EC (until 19 th April 2016)
6. Harmonised standards used:
BS EN61010-1:2010 third edition Safety requirements for electrical equipment for measurement, control, and laboratory use Part 1: General requirements
EN61326-1:2013 Electrical equipment for measurement, control and laboratory use – EMC requirements Part 1: General requirements
BS EN 60529:1992+A2:2013 Degrees of protection provided by enclosures (IP code)
7. Intertek Testing and Certification Ltd, No: 3272281, performed compliance testing to BS EN
61010-1:2010, IEC 61010-1:2010, UL 61010-1:2010 and CAN/CSA C22.2 Bo 6101010-
1:2010 and issued certification of compliance to these standards.
Signed for and behalf of:
Watson Marlow Ltd
Falmouth, 6 th January 2016
Simon Nicholson, Managing Director, Watson-Marlow Limited m-530sn-gb-02
2 Declaration of incorporation
Watson-Marlow Ltd
Falmouth
Cornwall
TR11 4RU
England
Declaration of Incorporation
In accordance with the Machinery Directive 2006/42/EC that if this unit is to be installed into a machine or is to be assembled with other machines for installations, it shall not be put into service until the relevant machinery has been declared in conformity.
We hereby declare that:
Peristaltic Pump
Series: 530 cased pumps the following harmonised standards have been applied and fulfilled for health and safety requirements:
Safety of Machinery – EN ISO 12100
Safety of Machinery – Electrical Equipment of Machines BS EN 60204-1
Quality Management System – ISO 9001 and the technical documentation is compiled in accordance with Annex VII(B) of the Directive.
We undertake to transmit, in response to a reasoned request by the appropriate national authorities, relevant information on the partly completed machinery identified above. The method of transmission shall be by mail or email.
The pump head is incomplete and must not be put into service until the machinery into which it is to be incorporated has been declared in conformity with the provisions of the Directive.
Person authorised to compile the technical documents:
Andrew Green, Design & Engineering Director, Watson-Marlow Ltd.
Place and date of declaration: Watson-Marlow Ltd, 31.07.2015
Responsible person:
Simon Nicholson, Managing Director, Watson-Marlow Ltd m-530sn-gb-02 5
3 When you unpack your pump
3.1
Unpacking your pump
Unpack all parts carefully, retaining the packaging until you are sure all components are present and in good order. Check against the components supplied list, below.
3.2
Packaging disposal
Dispose of packaging materials safely, and in accordance with regulations in your area. The outer carton is made of corrugated cardboard and can be recycled.
3.3
Inspection
Check that all components are present. Inspect components for damage in transit. If anything is missing or damaged, contact your distributor immediately.
3.4
Components supplied
530 Components
•
•
•
530 pump drive unit, fitted with 520R2 or other pumphead if specified as a pump
The designated power cable (attached to the pump drive unit)
A 520N module providing pump ingress protection to IP66, NEMA 4X, if a 530SN, 530UN, 530DuN or 530BpN.
Note: the module is attached for transit, but must be removed to allow wiring up, voltage selection and fuse inspection and then re-affixed before the pump is operated.
•
3.5
Product safety information booklet incorporating quick start manual
Storage
This product has an extended shelf life. However, care should be taken after storage to ensure that all parts function correctly. Please observe the storage recommendations and use-by dates which apply to tubing you may wish to bring into service after storage.
6 m-530sn-gb-02
4 Information for returning pumps
In compliance with the UK Health and Safety at Work Act and the Control of Substances Hazardous to Health
Regulations, you are required to declare the substances which have been in contact with product(s) you return to Watson-Marlow or it subsidiaries or distributors. Failure to do so will cause delays. Please ensure that you email us this information and receive a RGA (Returned Goods Authorisation) before you despatch the product(s). A copy of the RGA form must be attached to the outside of the packaging containing the product(s).
Please complete a separate decontamination certificate for each product and attach it to the outside of the packaging containing the product(s). A copy of the appropriate decontamination certificate can be downloaded from the Watson-Marlow website at www.wmftg.com
You are responsible for cleaning and decontaminating the product(s) before return.
m-530sn-gb-02 7
5 Peristaltic pumps - an overview
Peristaltic pumps are the simplest possible pump, with no valves, seals or glands to clog or corrode. The fluid contacts only the bore of a tube, eliminating the risk of the pump contaminating the fluid, or the fluid contaminating the pump. Peristaltic pumps can operate dry without risk.
How they work
A compressible tube is squeezed between a roller and a track on an arc of a circle, creating a seal at the point of contact. As the roller advances along the tube, the seal also advances. After the roller has passed, the tube returns to its original shape, creating a partial vacuum which is filled by fluid drawn from the inlet port.
Before the roller reaches the end of the track, a second roller compresses the tube at the start of the track, isolating a packet of fluid between the compression points. As the first roller leaves the track, the second continues to advance, expelling the packet of fluid through the pump’s discharge port. At the same time, a new partial vacuum is created behind the second roller into which more fluid is drawn from the inlet port.
Backflow and siphoning do not occur, and the pump effectively seals the tube when it is inactive. No valves are needed.
The principle may be demonstrated by squeezing a soft tube between thumb and finger and sliding it along: fluid is expelled from one end of the tube while more is drawn in at the other.
Animal digestive tracts function in a similar way.
Suitable applications
Peristaltic pumping is ideal for most fluids, including viscous, shear-sensitive, corrosive and abrasive fluids, and those containing suspended solids. They are especially useful for pumping operations where hygiene is important.
Peristaltic pumps operate on the positive displacement principle. They are particularly suitable for metering, dosing and dispensing applications. Pumps are easy to install, simple to operate and inexpensive to maintain.
8 m-530sn-gb-02
6 Warranty
Watson-Marlow Ltd ("Watson- Marlow") warrants this product to be free from defects in materials and workmanship for five years from the date of shipment, under normal use and service.
Watson-Marlow's sole responsibility and the customer's exclusive remedy for any claim arising out of the purchase of any product from Watson-Marlow is, at Watson-Marlow's option: repair, replacement or credit, where applicable.
Unless otherwise agreed in writing, the foregoing warranty is limited to the country in which the product is sold.
No employee, agent or representative of Watson- Marlow has the authority to bind Watson-Marlow to any warranty other than the foregoing unless in writing and signed by a director of Watson-Marlow. Watson-
Marlow makes no warranty of the fitness of its products for a particular purpose.
In no event: i.
shall the cost of the customer's exclusive remedy exceed the purchase price of the product; ii.
shall Watson- Marlow be liable for any special, indirect, incidental, consequential, or exemplary damages, however arising, even if Watson- Marlow has been advised of the possibility of such damages.
Watson-Marlow shall not be liable for any loss, damage, or expense directly or indirectly related to or arising out of the use of its products, including damage or injury caused to other products, machinery, buildings, or property. Watson- Marlow shall not be liable for consequential damages, including without limitation, lost profits, loss of time, inconvenience, loss of product pumped, and loss of production.
This warranty does not obligate Watson-Marlow to bear any costs of removal, installation, transportation, or other charges which may arise in connection with a warranty claim.
Watson-Marlow shall not be responsible for shipping damage of returned items.
Conditions
o o o o
Products must be returned by pre- arrangement to Watson- Marlow, or a Watson- Marlow approved service centre.
All repairs or modifications must have been made by Watson-Marlow Ltd, or a Watson-Marlow approved service centre or with the express permission in writing of Watson-Marlow, signed by a manager or director of Watson-Marlow.
Any remote control or system connections must be made in accordance to Watson- Marlow recommendations.
All PROFIBUS systems must be installed or certified by a PROFIBUS approved installation engineer.
Exceptions
o o o o
Consumable items including tubing and pumping elements are excluded.
Pumphead rollers are excluded.
Repairs or service necessitated by normal wear and tear or by lack of reasonable and proper maintenance are excluded.
Products which, in the judgement of Watson- Marlow, have been abused, misused, or subject to malicious or accidental damage or neglect are excluded.
o o o o
Failure caused by electrical surge is excluded.
Failure caused by incorrect or sub-standard system wiring is excluded.
Damage by chemical attack is excluded.
Ancillaries such as leak detectors are excluded.
o o
Failure caused by UV light or direct sunlight.
Any attempt to disassemble a Watson-Marlow product will invalidate the product warranty.
Watson-Marlow reserves the right to amend these terms and conditions at any time.
m-530sn-gb-02 9
7 Safety notes
This safety information should be used in conjunction with the rest of this operating manual.
In the interests of safety, this pump and pumphead should only be used by competent, suitably trained personnel after they have read and understood the manual and considered any hazard involved. If the pump is used in a manner not specified by Watson- Marlow Ltd, the protection provided by the pump may be impaired. Any person who is involved in the installation or maintenance of this equipment should be fully competent to carry out the work. In the UK this person should also be familiar with the Health and Safety at
Work Act 1974.
This symbol, used on the pump and in the manual, means: Caution, refer to accompanying documents.
This symbol, used on the pump and in the manual, means: Do not allow fingers to contact moving parts.
This symbol, used on the pump and in the manual, means: Caution, hot surface.
This symbol, used on the pump and in the manual, means: Caution, risk of electric shock.
This symbol, used on the pump and in the manual, means: Personal
Protective Equipment (PPE) must be worn.
This symbol, used on the pump and in the manual, means: Recycle this product under the terms of the EU Waste Electrical and Electronic
Equipment (WEEE) Directive.
Fundamental work with regard to lifting, transportation, installation, starting-up, maintenance and repair should be performed by qualified personnel only. The unit must be isolated from mains power while work is being carried out. The motor must be secured against accidental start-up.
10 m-530sn-gb-02
Some pumps weigh more than 18kg (the exact weight depends on the model and pumphead - see on the pump). Lifting should be performed according to standard Health and Safety guidelines. Finger recesses are built into the sides of the lower shell for convenience in lifting; in addition, the pump can conveniently be lifted by grasping the pumphead and (where fitted) the 'N' module at the rear of the pump.
There is a user-replaceable type fuse in the fuse holder in the centre of the switch-plate at the back of the pump. In some countries, the mains power plug contains an additional replaceable fuse.
There are no user-serviceable fuses or parts inside this pump.
This pump must be used only for its intended purpose.
The pump must be accessible at all times to facilitate operation and maintenance. Access points must not be obstructed or blocked. Do not fit any devices to the drive unit other than those tested and approved by
Watson-Marlow. Doing so could lead to injury to persons or damage to property for which no liability can be accepted.
The pump's main plug is the disconnecting device (for isolating the motor drive from the mains supply in an emergency). Do not position the pump so that it is difficult to disconnect the mains plug.
If hazardous fluids are to be pumped, safety procedures specific to the particular fluid and application must be put in place to protect against injury to persons.
This product does not comply with the ATEX directive and must not be used in explosive atmospheres.
Ensure the chemicals to be pumped are compatible with the pumphead, lubricant (where applicable), tubing, pipework and fittings to be used with the pump. Please refer to the chemical compatibility guide which can be found at: www.wmftg.com/chemical .
If you need to use the pump with any other chemical please contact
Watson-Marlow to confirm compatibility.
m-530sn-gb-02 11
Pump will start as soon as power is applied if Auto restart is on and pump was running when power was disconnected.
There are moving parts inside the pumphead. Before opening the toolunlockable guard or tool-unlockable track, ensure that the following safety directions are followed:
1.
2.
3.
Ensure the pump is isolated from the mains power.
Ensure that there is no pressure in the pipeline.
4.
If a tube failure has occurred, ensure that any fluid in the pumphead has been allowed to drain to a suitable vessel, container or drain.
Ensure that appropriate Personal Protective Equipment (PPE) is worn.
Primary operator protection from rotating parts of the pump is provided by the pumphead safeguard. Note that safeguards differ, depending on the type of pumphead. See the pumphead section of the manual.
12 m-530sn-gb-02
8
8.1
Pump specifications
Specification ratings
Operating temperature
Storage temperature
Humidity (non-condensing)
Maximum altitude
Power consumption
Supply voltage
Maximum voltage fluctuation
Full load current
Fuse rating
Installation category
(overvoltage category)
Pollution degree
IP dB rating
5C to 40C (41F to 104F)
530: -40C to 70C (-40F to 158F)
80% up to 31C (88F) decreasing linearly to 50% at 40C (104F)
2000m (6560ft)
530: 135VA
Filtered 100-120V/200-240V 50/60Hz 1pH
+/-10% of nominal voltage. A well regulated electrical mains supply is required along with cable connections conforming to the best practice of noise immunity
530: <0.6A@ 230V; <1.25A @ 115V
530: Ceramic, 5x20mm, 2.5A, 250V AC, Time Delay
II
2
IP31 to BS EN 60529 if supplied with N module then IP66 to BS EN
60529. Equivalent to NEMA 4X to NEMA 250 *(indoor use - protect from prolonged UV exposure)
530: < 70dB (A) @ 1m
Control ratio
Maximum speed
8.2
Weights
530: 0.1-220rpm (2200:1)
530: 220rpm
530
IP31
IP66
Drive only
9.7kg
10.6kg
21lb 6oz
23lb 5oz
+ 520R, 520R2 + 520REL,520REM,
520REH
10.6kg
23lb 5oz 10.5kg
23lb 3oz
11.5kg
25lb 5oz 11.4kg
25lb 2oz
+ 505L
12.2kg
13.1kg
26lb 14oz
28lb 13oz
Weights warning
Some pumps weigh more than 18kg (the exact weight depends on the model and pumphead - see on the pump). Lifting should be performed according to standard Health and Safety guidelines. Finger recesses are built into the sides of the lower shell for convenience in lifting; in addition, the pump can conveniently be lifted by grasping the pumphead and (where fitted) the 'N' module at the rear of the pump.
m-530sn-gb-02 13
8.3
Pumphead options
530 pump range
520R, 501RL, 313, 314, 505L, 505BA, 505CA, 314MC, 318MC:
14 m-530sn-gb-02
9 Good pump installation practice
9.1
General recommendations
It is recommended that the pump is sited on a flat, horizontal, rigid surface, free from excessive vibration, to ensure the correct lubrication of the gearbox and correct pumphead operation. Allow a free flow of air around the pump to ensure that heat can be dissipated. Ensure that the ambient temperature around the pump does not exceed the recommended maximum operating temperature.
The STOP key on pumps supplied with a keypad will always stop the pump. However, it is recommended that a suitable local emergency stop device is fitted into the mains supply to the pump.
Do not stack the pumps more than the recommended maximum number. When the pumps are stacked, ensure that the ambient temperature around all the pumps in the stack does not exceed the recommended maximum operating temperature.
The pump may be set up so that the direction of rotor rotation is clockwise or counter-clockwise, whichever is convenient.
Please note, however, that for some pumpheads the tube life will be greater if the rotor rotates clockwise; and that performance against pressure will be maximised if the rotor rotates counter-clockwise. To achieve pressure in some pumpheads the pump must rotate counter-clockwise.
Peristaltic pumps are self- priming and self- sealing against backflow. No valves are required in inlet or discharge line, except those specified as below.
Users must fit a non-return valve between the pump and the discharge pipework to avoid the sudden release of pressurised fluid in the event of a pumphead or tube failure. This shall be fitted immediately after the discharge of the pump.
Valves in the process flow must be opened before the pump operates. Users are advised to fit a pressure relief device between the pump and any valve on the discharge side of the pump to protect against damage caused by accidental operation with the discharge valve closed.
m-530sn-gb-02 15
9.2
Dos and don'ts
l l l l l l l l l
Do not build a pump into a tight location without adequate airflow around the pump.
Do keep delivery and suction tubes as short and direct as possible - though ideally not shorter than one metre - and follow the straightest route. Use bends of large radius: at least four times the tubing diameter. Ensure that connecting pipework and fittings are suitably rated to handle the predicted pipeline pressure. Avoid pipe reducers and lengths of smaller bore tubing than the pumphead section, particularly in pipelines on the suction side. Any valves in the pipeline must not restrict the flow. Any valves in the flow line must be open when the pump is running.
Do ensure that on longer tube runs at least one metre of smooth bore, flexible tubing is connected to the inlet and discharge port of the pumphead to help to minimise impulse losses and pulsation in the pipeline.
This is especially important with viscous fluids and when connecting to rigid pipework.
Do use suction and delivery pipes equal to or larger than the tubing diameter bore. When pumping viscous fluids use pipe runs with a bore several times larger than the pump tube.
Do site the pump at or just below the level of the fluid to be pumped if possible. This will ensure flooded suction and maximum pumping efficiency.
Do run at slow speed when pumping viscous fluids. Flooded suction will enhance pumping performance, particularly for materials of a viscous nature.
Do recalibrate after changing tubing, fluid or any connecting pipework. It is also recommended that the pump is recalibrated periodically to maintain accuracy.
Do not pump any chemical not compatible with the tube or pumphead.
Do not run the pump with no tube or element fitted to the pumphead.
l l l l
Do not strap the control and mains cable together.
Do ensure if your product has an N module, that the module is fitted with the seals intact and properly located. Ensure that the holes for the cable glands are properly sealed to maintain the IP/NEMA rating.
If your pump is a PROFIBUS variant do ensure that it is installed in accordance with PROFIBUS installation guidelines.
Do not allow tight bends in any PROFIBUS signal cable.
Tube selection: The chemical compatibility guide published on the Watson Marlow website is for guidance. If in any doubt about the compatibility of a tube material and the duty fluid, request a Watson- Marlow tube sample card for immersion trials.
When using Marprene or Bioprene continuous tubing, do re-tension the tube after the first 30 minutes of running.
16 m-530sn-gb-02
10 Pump operation
10.1
Keypad Layout and Key IDs
HOME key
When the HOME key is pressed it will return the user to the last known operating mode. If modifying pump settings when the HOME key is pressed, it will disregard any setting changes and return you to the last known operating mode.
FUNCTION keys
FUNCTION keys, when pressed, will perform the function displayed on the screen directly above the relevant function key.
∧ and
∨ keys
These keys are used to change the programmable values within the pump. These keys are also used to move the selection bar up and down in the menus.
MODE key
To change modes or mode settings, press the MODE key. The MODE key can be pressed at any time to enter the mode menu. If modifying pump settings when the MODE key is pressed, it will disregard any setting changes and return you to the MODE menu.
m-530sn-gb-02 17
10.2
Starting and stopping
10.3
Using up and down keys
10.4
Maximum speed
10.5
Change rotation direction
18 m-530sn-gb-02
11 Connecting to a power supply
A well regulated electrical mains supply is required along with cable connections conforming to the best practice of noise immunity. It is not recommended to site these drives alongside 'dirty' electrical mains devices, such as 3- phase contactors and inductive heaters without special attention being paid to unacceptable mains-borne noise.
Set the voltage selector to 115V for 100-120V 50/60Hz supplies or
230V for 200-240V 50/60Hz supplies. Always check the voltage selector switch before connecting to the mains supply or the pump will be damaged.
Make suitable connection to an earthed single-phase mains electricity supply.
If the pump type is one with an 'N' module, the voltage selector is not visible while the module is in place. It is mounted in the switch plate at the rear of the pump, protected from water by the 'N' module. The module must be removed to allow access to the switch plate. Do not switch the pump on unless you have checked that it is set to suit your power supply by removing the module and inspecting the switch and then refitting the module.
1.
4.
7.
2.
5.
8.
3.
6.
9.
10.
We recommend using a commercially available supply voltage surge suppression where there is excessive electrical noise.
m-530sn-gb-02 19
Ensure that all power supply cables are adequately rated for the equipment.
The pump must be positioned so that the disconnection device is easily accessible when the equipment is in use.
11.1
Conductor colour coding
Conductor type
Line
Neutral
Ground
European colour
Brown
Blue
Green/Yellow
North American colour
Black
White
Green
20 m-530sn-gb-02
12 Start-up check list
Note : See also Tube loading.
o o o
Ensure that proper connections are achieved between the pump and suction and discharge piping.
Ensure proper connection has been made to a suitable power supply.
Ensure that the recommendations in the section are followed.
"Good pump installation practice" on page 15
m-530sn-gb-02 21
13 Switching the pump on for the first time
Power up the pump. The pump displays the start-up screen with the Watson-Marlow
Pumps logo for three seconds.
13.1
Selecting the display language
1. Use the and press
∧
/
∨ keys to select your chosen language,
SELECT .
2. Your selected language will now be displayed on sceen. Choose CONFIRM to continue. All text will now appear in your chosen language.
3. Choose REJECT to return to the language choice screen. This then proceeds to the home screen.
22 m-530sn-gb-02
13.2
First-time start-up defaults
The pump is preset with operational parameters as shown in table below.
Parameter
Language
Default mode
Default manual speed
Pump status
Max speed
Direction
Pumphead
Tube size
Tube material
Flow calibration
Flow units
SG value
Keypad lock
Auto-restart
Analog signal type
Analog scaling type
Analog min current
Analog max current
Analog min flow rate/rpm
Analog max flow rate/rpm
Beeper
Security code
MemoDose flow rate
MemoDose volume
Pump number baud rate
Stop bits
Profibus node address
Remote start/stop input
Leak detector input
Input 4
Input 5
Output 1
Output 1 - status
530 default
Not set
Manual
220rpm
Stopped
220rpm
CW
520R2
9.6mm
Bioprene
15.12 ml/rev rpm
1
Disabled
OFF mA mA
5mA
19mA
0rpm
220rpm
ON
Not set
Mid flow rate of chosen pumphead
100ml
9600
2
126
High = stop
High = leak
Disabled
Disabled
Run/stop
High = run m-530sn-gb-02 23
Parameter
Output 2
Output 2 - status
Output 3
Output 3 - status
Output 4
Output 4 - status
Direction
High = CW
Auto/man
High = auto
General alarm
High = alarm
530 default
The pump is now ready to operate according to the defaults listed above.
Note: The display background colour changes according to running state as follows: l l
White background indicates pump stopped
Grey background indicates pump running l
Red background indicates error or alarm
All operating parameters may be changed by means of key-presses (see section
).
24 m-530sn-gb-02
14 Switching the pump on in subsequent power cycles
Subsequent power-up sequences will jump from the start-up screen to the home screen.
l l l
The pump runs a power-on test to confirm proper functioning of the memory and hardware. If a fault is found, an error code is displayed.
The pump displays the start-up screen with the Watson-Marlow Pumps logo for three seconds followed by the home screen
Start-up defaults are those in place when the pump was switched off last
Check that the pump is set to operate as you require it. The pump is now ready to operate.
All operating parameters may be changed by means of key- presses (see
Power interruption
This pump has an auto restart feature which, when active, will restore the pump to the operating state it was in when power was lost.
Stop/start power cycles
Do not power up/power down the pump more than 20 times in an hour, whether manually or by means of the auto-restart facility. We recommend remote control where a high frequency of stop/start cycles is required.
Pump will start as soon as power is applied if Auto restart is on and pump was running when power was disconnected.
m-530sn-gb-02 25
15 Mode menu
Press MODE to display the Change mode menu.
Use the
∧ and
∨ keys to scroll through the available modes: o o o o
Manual (default)
Flow calibration
MemoDose
CANCEL
Use SELECT to choose mode. Use the right hand function key to alter mode settings.
26 m-530sn-gb-02
16 Manual
All settings and functions of the pump in manual mode are set and controlled by means of key- presses.
Immediately after the start- up display sequence detailed in:
"Switching the pump on in subsequent power cycles" on page 25
, the manual mode home screen will be displayed unless auto restart is enabled.
If enabled, the pump will return to the last known operating state when the power was lost. When the pump is running it displays an animated clockwise arrow. In normal operation, the direction of flow is into the bottom port of the pumphead and out of the top port.
If an exclamation mark (!) is displayed, it indicates that Auto restart is on (see
41 ). If a padlock icon shows, it indicates that the keypad lock is on.
16.1
START
Starts the pump at the current flow displayed, and the display background changes to grey. If the pump is running, pressing this has no effect.
m-530sn-gb-02 27
16.2
STOP
Stops the pump. The display background changes to white. If the pump is not running pressing this has no effect.
16.3
INCREASING AND DECREASING FLOW RATE
Using the
∧ and
∨ keys will increase or decrease the flow rate.
Decreasing flow rate
o o o
A single key press will decrease flow rate by the least significant digit of the chosen flow rate unit.
Repeat key presses as required to achieve the desired flow rate.
Hold down the key for flow rate scrolling.
28 m-530sn-gb-02
Increasing flow rate
o o o
A single key press will increase flow rate by the least significant digit of the chosen flow rate unit.
Repeat key presses as required to achieve the desired flow rate.
Hold down the key for flow rate scrolling.
16.4
MAX FUNCTION (Manual mode only)
o o o Press and hold the MAX key to run at maximum flow.
Release the key to stop the pump.
The volume dispensed and time elapsed are displayed while the MAX key is pressed and held.
m-530sn-gb-02 29
17 Flow calibration
This pump displays flow rate in ml/min.
17.1
Setting the flow calibration
Using the
∧
/
∨ keys, scroll to Flow calibration and press CALIBRATE .
Using the
∧
/
∨ keys, enter the maximum flow rate limit and press ENTER .
30 m-530sn-gb-02
Press START to begin pumping a volume of fluid for calibration.
Press STOP to stop pumping fluid for the calibration.
m-530sn-gb-02 31
Use the
∧
/
∨ keys to enter the actual volume of fluid pumped.
To accept the new calibration press or MODE to abort.
ACCEPT or RE-CALIBRATE to repeat the procedure. Press HOME
The pump is now calibrated.
32 m-530sn-gb-02
18 MemoDose mode
Each time the pump is started by pressing START , it records the number of pumphead revolutions which occur until STOP is pressed. The number of revolutions is proportional to the volume of fluid which has been dispensed: the dose. MemoDose mode allows the user to repeat-dose a precise volume of fluid. This can be achieved by dispensing a master dose, or entering a dose volume manually using the keypad. MemoDose can repeat this dose exactly or proportionately.
18.1
To configure MemoDose
o o
Select MODE
Using the
∧
/
∨ keys, scroll to MemoDose and press SETTINGS
Note: the pump must be stopped to enter MemoDose Settings.
m-530sn-gb-02 33
18.2
Set flow rate
Using the
∧
/
∨ keys, scroll to Flow rate and press SELECT .
Using the
∧
/
∨ keys, enter the dose flow rate and press SELECT .
34 m-530sn-gb-02
18.3
Resume interrupted doses
MemoDose Mode gives you the opportunity to resume interrupted doses following a power cycle (Note:
Auto-restart must be on for this to function). Alternatively interrupted doses can be discarded and a new dose started when power is resumed.
From the MemoDose settings screen, use the press
∧
/
∨ keys to scroll to Resume Interrupted Dose and
ENABLE . The red cross will become a green tick to indicate that Resume interrupted dose is enabled.
If this function is enabled the ENABLE key becomes DISABLE . If this is pressed interrupted doses will be discarded after a power cycle.
If resume interrupted doses has been set the following warning screen is displayed, press save this setting.
CONFIRM to m-530sn-gb-02 35
If resume interrupted doses has NOT been set the following warning screen is displayed, press to save this setting.
CONFIRM
18.4
Master dose
Using the
∧
/
∨ keys, scroll to Master dose and press SELECT .
36 m-530sn-gb-02
The following screen is displayed; press master dose.
MANUAL to enter a dose via the keypad or DOSE to dispense a
Dispensing a master dose
Press START to begin dispensing the Master Dose.
m-530sn-gb-02 37
Press STOP to finish dispensing the Master Dose.
Manual dose entry
Press MANUAL on the master dose screen. Using the
∧
/
∨ keys enter the required dose volume and press
SELECT to record the master dose or CANCEL to return to MemoDose Settings.
38 m-530sn-gb-02
Save dose volume
Press SAVE to record the Master Dose or CANCEL to return to MemoDose settings.
When MemoDose SETTINGS are complete the following screen is displayed; press
MemoDose Mode or BACK to return to MemoDose settings.
MEMODOSE to start in m-530sn-gb-02 39
18.5
Manual dosing
From the MemoDose Home screen, press START to deliver a dose. The screen displays the dose flow rate and the dose remaining counting down from 100% to 0%.
If the delivered dose differs from the volume required, the percentage may be adjusted within the limits 1% to 999% of the master dose. Use the
∧
/
∨ keys to alter the percentage. The new dose size is displayed as a percentage on the home screen.
If STOP is pressed during dosing, the pump stops. Pressing START will resume, or discard, the interrupted dose depending on the settings in
40 m-530sn-gb-02
19 Main menu
To access the main menu press the MENU button from one of the HOME screens or INFO screens.
This will display the main menu as shown below. Use the available options.
∧
/
∨ keys to move the selection bar between the
Press SELECT to choose an option.
Press EXIT to return to the screen from where the MENU was called.
m-530sn-gb-02 41
19.1
Security settings
Security settings can be changed by selecting SECURITY SETTINGS from the Main menu.
Auto keypad lock
Press ENABLE / DISABLE to switch on/off the Auto keypad lock. When active the keypad will 'lock' after
20 seconds of inactivity.
Once locked it will display the screen below when any key is pressed. To unlock the keypad press the two
UNLOCK keys together.
42 m-530sn-gb-02
The padlock icon will appear on the operating mode home screen to show that keypad lock is activated.
Note that the STOP key will always work whether the keypad is locked or not.
PIN protection
Using the
∧
/
∨ keys select PIN protection from the SECURITY SETTINGS menu and press
ENABLE/DISABLE to switch on/off the PIN protection. If PIN protection has been enabled, a Master level
PIN will be required to disable PIN lock.
Setting Master PIN
Setting the Master PIN protects all functionality. The Master is able to selectively enable functionality for two additional operators. These are defined as User 1 and User 2. They will be able to access this functionality by entering a PIN code assigned to them by the Master user. To set the Master PIN, scroll to Master level and press ENABLE .
m-530sn-gb-02 43
To define a four digit Master PIN, use the digit press the
∧
/
∨ keys to select each digit from 0-9. Once you have the required
NEXT DIGIT key. After selecting the fourth digit press ENTER .
Now press CONFIRM to PIN entry.
to check that the number entered is the PIN you require. Press CHANGE to return
44 m-530sn-gb-02
The following screen will be displayed to indicate that the Master PIN has been applied to access all functionality. Press NEXT to selectively enable functionality access for User 1 and User 2.
Configure User 1 security settings
The PIN PROTECTION level screen will be displayed with User 1 highlighted, press
User 1 security settings or scroll to configure an alternative User.
ENABLE to configure m-530sn-gb-02 45
ENABLE user 1 security settings displays the PIN entry screen for User 1. To define a four digit User 1 PIN, use the
∧
/
∨ keys to select each digit from 0-9. Once you have the required digit press the NEXT DIGIT key. After selecting the fourth digit press ENTER .
Now press CONFIRM to PIN entry.
to verify that the number entered is the PIN you require. Press CHANGE to return
46 m-530sn-gb-02
To define the allowed functionality, use the
∧
/
∨ keys to select the functionality and press ENABLE . User 1
PIN will allow access to only the enabled functionality, to disable functionality, highlight the enabled functionality and press DISABLE . When all the required functionality has been enabled, press FINISH .
Configure User 2 security settings
The PIN PROTECTION level screen will be displayed with User 2 highlighted, press
User 2 security settings or scroll to configure an alternative User.
ENABLE to configure m-530sn-gb-02 47
ENABLE user 2 security settings displays the PIN entry screen for User 2. To define a four digit User 2 PIN, use the
∧
/
∨ keys to select each digit from 0-9. Once you have the required digit press the NEXT DIGIT key. After selecting the fourth digit press ENTER .
To define the allowed functionality, use the
∧
/
∨ keys to select the functionality and press ENABLE . User 2
PIN will allow access to only the enabled functionality, to disable functionality, highlight the enabled functionality and press DISABLE . When all the required functionality has been enabled, press FINISH .
48 m-530sn-gb-02
Note : Once Security Settings for User 1 and User 2 have been set by the Master, only the Master PIN will allow access to Security Settings.
The HOME screen will be displayed. A PIN is now required to access all functionality. The Master PIN accesses all pump functionality and the User 1 and User 2 PINs access only the defined functionality. To enter the PIN, use the
∧
/
∨ keys to select each digit from 0-9. Once you have the required digit press the
NEXT DIGIT key. After selecting the fourth digit press ENTER .
If an incorrect PIN has been entered the following screen will be displayed. NOTE: this screen will also display if the PIN entered does not allow access to that functionality.
If a PIN number is entered that is already in use, the following screen will be displayed, press input an alternative PIN or EXIT to abort
CHANGE to m-530sn-gb-02 49
If the PIN entered does not allow access to the functionality the following screen will be displayed.
Keypad beep
From SECURITY settings scroll to Keypad beep using the beep at every key press.
∧
/
∨ keys and select ENABLE . The pump will now
50 m-530sn-gb-02
19.2
General settings
To view the general settings menu, select GENERAL SETTINGS from the main menu.
Auto restart
This pump offers an auto restart feature. If active on power loss, it restores the pump when power returns to the operating state it was in when power was lost.
For example, if the pump was running in analog mode prior to power loss, it would return to the same operating mode and continue to run at a proportional speed to the analog input.
Press ENABLE / DISABLE to turn the auto restart feature on/off.
Do not use auto restart for more than 20 mains power starts per hour.
We recommend remote control where a high number of starts is required.
Pump will start automatically if start conditions are met.
m-530sn-gb-02 51
The !
icon is displayed on the home screens to indicate that the auto restart feature is active.
Flow units
The current chosen flow unit is displayed on the right hand side of the screen. To change flow units move the selection bar over the flow unit menu entry and press SELECT .
Use the
∧
/
∨ keys to move the selection bar over the required flow unit, then press SELECT . All flow rates displayed on screens will now be in the chosen units.
52 m-530sn-gb-02
If a mass flow unit is selected, the specific gravity of the fluid must be entered. The following screen is displayed.
Use the
∧
/
∨ keys to enter the value of the specific gravity, and press SELECT .
Pump label
The pump label is a user defined 20 digit alphanumeric label which is displayed in the header bar of the home screen. To define or edit the pump label, move the selection bar over the Pump label menu entry and press
SELECT . If a pump label has been previously defined, this will be displayed on screen to allow editing, otherwise it will display the default label “WATSON-MARLOW".
m-530sn-gb-02 53
Use the
∧
/
∨ keys to scroll through the available characters for each digit. The available characters are 0- 9,
A-Z and SPACE.
Press NEXT to move onto the next character, or PREVIOUS to move back to the previous character.
Press FINISH to save the entry and return to the general settings menu.
54 m-530sn-gb-02
Pumphead type
Select GENERAL SETTINGS from the main menu.
Use the
∧
/
∨ keys to move the selection bar over Pumphead type and press SELECT . The following screen will be displayed.
Use the
∧
/
∨ keys to move the selection bar over Pumphead and press SELECT .
m-530sn-gb-02 55
Use the
∧
/
∨ keys to move the selection bar over the required pumphead type and press SELECT .
Tube size and tube material
Select Tube size from GENERAL SETTINGS, then use the
∧
/
∨ keys to move the selection bar over Bore size and press SELECT .
56 m-530sn-gb-02
Use the
∧
/
∨ keys to move the selection bar over the tube size to be used and press SELECT .
If a LoadSure element has been selected then the tube size is displayed as pressure and bore.
m-530sn-gb-02 57
This screen also allows you to select the tube material used.
Use the
∧
/
∨ keys to move the selection bar over Tube material and press SELECT .
Use the
∧
/
∨ keys to move the selection bar over the tube material to be used and press SELECT .
58 m-530sn-gb-02
The PUMPHEAD MODEL screen allows the tube Lot Number to be recorded for future reference. Use the
∧
/
∨ keys to move the selection bar over Tube lot number and press SELECT .
Use the
∧
/
∨ keys to scroll through the available characters for each digit. The available characters are 0- 9,
A-Z, and SPACE.
Press NEXT to move onto the next character, or PREVIOUS to move back to the last character.
Press FINISH to save the entry and return to the general settings menu.
Restore defaults
To restore the factory default settings select Restore defaults from the general settings menu.
There are two confirmation screens to ensure that this function is not carried out in error.
Press CONFIRM followed by RE-CONFIRM to restore the defaults.
m-530sn-gb-02 59
Language
Select language from the general settings menu to choose an alternative display language for the pump. The pump must be stopped before changing the language.
Use the
∧
/
∨ keys to move the selection bar to your required language. Press SELECT to confirm.
Your selected language will now be displayed on screen. Press CONFIRM to continue, all displayed text will now appear in your chosen language.
Press REJECT to return to the language choice screen.
MODE menu
Selecting MODE menu from the main menu will navigate you to access the sub- menu shown below. This is the same as pressing the MODE key. Please see
for further details.
60 m-530sn-gb-02
19.3
Control settings
Select CONTROL SETTINGS from the main menu to access the sub menu shown below. Use the
∧
/
∨ keys to move the selection bar. Press SELECT to choose the required function.
Speed limit
The maximum speed the pump is capable of running at is 220rpm.
Select Speed limit from the control settings menu to define a lower maximum speed limit for the pump.
This speed limit will be applied to all operating modes.
Use the
∧
/
∨ keys to adjust the value and press SAVE to set.
m-530sn-gb-02 61
Reset run hours
Select Reset run hours from the control settings menu.
Select RESET to zero the run hours counter. The run hours counter can be viewed by pressing INFO from your home screen. The following screen will be displayed. Press RESET to reset the run hours or CANCEL to return to the CONTROL SETTINGS menu.
19.4
Help
Select Help from the main menu to access the help screens.
62 m-530sn-gb-02
20 Troubleshooting
If the pump display remains blank when the pump is switched on, make the following checks:
•
•
•
•
•
Check that the mains power is available to the pump.
Check the fuse in the wall plug if one is present.
Check the position of the voltage selector switch.
Check the mains power switch at the rear of the pump.
Check the fuse in the fuse holder in the centre of the switch plate at the rear of the pump.
If the pump runs but there is little or no flow, make the following checks:
•
•
•
•
•
•
•
•
Check that fluid is supplied to the pump.
Check for any kinks or blockages in the lines.
Check that any valves in the line are open.
Check that the tube and rotor are in the pumphead.
Check that the tube is not split or burst.
Check that the correct wall-thickness tube is being used.
Check the direction of rotation.
Check that the rotor is not slipping on the drive shaft.
If the pump turns on, but will not run:
•
•
•
Check the remote stop function and configuration.
Check the mode you are in, are you in analog mode.
Try to operate and run the pump in manual mode.
20.1
Error codes
If an internal error occurs, an error screen with a red background is displayed. Note: Signal out of range, over signal and leak detected error screens report the nature of an external condition. They do not flash.
Error code Error condition
Er 0
Er 1
Er 2
Er 3
Er 4
Er 9
FRAM write error
FRAM corruption
FLASH write error during drive update
FLASH corruption
FRAM shadow error
Motor stalled
Suggested action
Attempt to reset by switching power OFF/ON. Or seek support.
Attempt to reset by switching power OFF/ON. Or seek support.
Attempt to reset by switching power OFF/ON. Or seek support.
Attempt to reset by switching power OFF/ON. Or seek support.
Attempt to reset by switching power OFF/ON. Or seek support.
Stop pump immediately. Check pumphead and tube.
Power OFF/ON may reset. Or seek support.
m-530sn-gb-02 63
Er15
Er16
Er17
Er19
Er20
Er21
Error code
Er10
Error condition
Tacho fault
Er14 Speed error
Er30
Over current
Over voltage
Under voltage
Over temperature
Signal out of range
Over signal
Over power
Err50 Communication error
20.2
Technical support
Watson-Marlow Fluid Technology Group
Falmouth, Cornwall
TR11 4RU
UK
Telephone: +44 (0) 1326 370370
Fax: +44 (0) 1326 376009
Email: [email protected]
www.wmftg.com
Suggested action
Stop pump immediately.
Power OFF/ON may reset. Or seek support.
Stop pump immediately.
Power OFF/ON may reset. Or seek support.
Stop pump immediately.
Power OFF/ON may reset. Or seek support.
Stop pump immediately. Check supply.
Power OFF/ON may reset.
Stop pump immediately. Check supply.
Power OFF/ON may reset.
Stop pump immediately. Turn OFF. Seek support.
Check analog control signal range. Trim signal as required. Or seek support.
Reduce the analog control signal.
Turn OFF. Check power supply. Check pumphead and tubing. Wait 30 minutes. Power ON may reset.
Or seek support.
Attempt to reset by switching power OFF/ON. Or seek support.
64 m-530sn-gb-02
21 Drive maintenance
There are no user serviceable parts inside the pump. The unit should be returned to Watson- Marlow for service.
m-530sn-gb-02 65
22 Drive spares
Description
Replaceable main fuse, type T2, 2.5A H 250V 20mm
Foot
Module seal
Module PCB(SCADA)
Module PCB (NEMA)
Module switch cover
Glands (STD)
Glands (EMC)
Blanking plugs
Sealing washer for blanking plug and gland
Snap-fit vent
Part No.
FS0064
MN2507M
MN2516B
MNA2066A
MNA2069A
MN2505M
GR0056
GR0075
GR0057
GR0058
MN2513B
66 m-530sn-gb-02
23 Pumphead replacement
Always isolate the pump from the mains power supply before opening any guard or track, or performing any positioning, removal or maintenance operation.
23.1
Pumphead replacement
520R
1.
2.
3.
4.
5.
6.
m-530sn-gb-02 67
1.
4.
24 Tube replacement
Always isolate the pump from the mains power supply before opening any guard or track, or performing any positioning, removal or maintenance operation.
24.1
Continuous tubing
520R and 520R2
2.
5.
3.
6.
7.
505L
8.
1.
2.
4.
5.
≤8.0mm=145mm ,
9.6mm=150mm
68
9.
3.
6.
m-530sn-gb-02
24.2
Tube elements
Grey
>2 bar (30psi)
Beige
>4 bar (60psi)
Blue
>7 bar (100psi)
520REL, 520REM and 520REH
1.
2.
4.
Sanitary connectors
5.
1.
Industrial connectors
2.
1.
2.
m-530sn-gb-02
3.
3.
3.
69
505L
1.
2.
3.
4.
General guide to cleaning with solvents
Chemical
Aliphatic hydrocarbons
Aromatic hydrocarbons
Ketone solvents
Halogenated/chlorinated solvents
Alcohols, general
Glycols
Estersolvents
Ether solvents
Cleaning precautions
Remove guard. Minimize rotor cap and clutch boot exposure to less than one minute (risk of attack).
Remove guard. Minimize rotor cap and clutch boot exposure to less than one minute (risk of attack).
Remove guard. Minimize rotor cap and clutch boot exposure to less than one minute (risk of attack).
Not recommended: possible risk to polycarbonate tube clamp adjusters and polypropylene tube clamp locators.
No precaution necessary.
Minimize rotor cap and clutch boot exposure to less than one minute (risk of attack).
Remove guard. Minimize rotor cap and tube clamp location cap exposure to less than one minute (risk of attack).
Not recommended: possible risk to polycarbonate tube clamp adjusters and polypropylene tube clamp locators.
70 m-530sn-gb-02
25 Ordering information
25.1
Pump part numbers
530 Part Code Configurator
0 5 ● 9 1 ● ● ● ● ●
Configuration
0: Pump
6: Drive only
Model
3: S
4: U
5: Du
6: Bp
Ingress
Protection*
1: IP31
N: IP66
*IP66 pumps are supplied with no plug or an
American mains plug.
Pumphead
EL: 520REL
EM: 520REM
EH: 520REH
2L: 520R2
10: 520R
5L: 505L
Plug options*
0: no plug
U: UK mains plug
E: EU mains plug
A: American mains plug
K: Australia mains plug
R: Argentina mains plug
C: Swiss mains plug
D
:
India/South
Africa mains plug
m-530sn-gb-02 71
25.2
Tubing and element part numbers
1.6mm wall tubing for 520R pumpheads
mm inch # Marprene Bioprene
4.8
6.4
8.0
mm
0.8
0.5
0.8
1.6
3.2
1.6
3.2
4.8
6.4
8.0
mm
0.5
0.8
1.6
3.2
4.8
6.4
8.0
1/50
1/32
1/16
1/8
3/16
1/4
5/16 inch
1/32
1/16
1/8
3/16
1/4
5/16 inch
1/50
1/32
1/16
1/8
3/16
1/4
5/16
112
13
14
16
25
17
18
#
13
14
16
25
17
18
#
112
13
14
16
25
17
18
902.0005.016
902.0008.016
902.0016.016
902.0032.016
902.0048.016
902.0064.016
902.0080.016
PVC
—
950.0016.016
950.0032.016
950.0048.016
950.0064.016
950.0080.016
Pumpsil
913.A005.016
913.A008.016
913.A016.016
913.A032.016
913.A048.016
913.A064.016
913.A080.016
933.0005.016
933.0008.016
933.0016.016
933.0032.016
933.0048.016
933.0064.016
933.0080.016
Fluorel
—
970.0016.016
970.0032.016
970.0048.016
970.0064.016
970.0080.016
STA-PURE
Series PCS
—
—
960.0016.016
960.0032.016
960.0048.016
960.0064.016
960.0080.016
Note: 1.6mm wall STA-PURE Series PFL and STA-PURE Series PCS tubing are supplied in 305mm lengths.
STA-PURE
Series PFL
—
—
965.0016.016
965.0032.016
965.0048.016
965.0064.016
965.0080.016
Neoprene
920.0008.016
920.0016.016
920.0032.016
920.0048.016
920.0064.016
920.0080.016
72 m-530sn-gb-02
2.4mm wall tubing for 520R2 pumpheads
mm
0.5
0.8
1.6
3.2
4.8
6.4
8.0
9.6
mm
0.8
1.6
3.2
4.8
6.4
8.0
inch
1/50
1/32
1/16
1/8
3/16
1/4
5/16
3/8 inch
1/32
1/16
1/8
3/16
1/4
5/16
#
—
—
119
120
15
24
121
122
#
—
119
120
15
24
121
Marprene
—
—
902.0016.024
902.0032.024
902.0048.024
902.0064.024
902.0080.024
902.0096.024
STA-PURE
Series PFL
—
965.0016.024
965.0032.024
965.0048.024
965.0064.024
965.0080.024
Bioprene
—
—
933.0016.024
933.0032.024
933.0048.024
933.0064.024
933.0080.024
933.0096.024
STA-PURE
Series PCS
—
960.0016.024
960.0032.024
960.0048.024
960.0064.024
960.0080.024
Pumpsil
913.A005.024
913.A008.024
913.A016.024
913.A032.024
913.A048.024
913.A064.024
913.A080.024
913.A096.024
Note: 2.4mm wall STA-PURE Series PFL and STA-PURE Series PCS tubing are supplied in 355mm lengths.
m-530sn-gb-02 73
2.4mm wall elements for 520RE pumpheads
0-2 bar (0-30 psi) pressure rated elements
Industrial mm
3.2
6.4
9.6
Sanitary mm
3.2
6.4
9.6
inch
1/8
1/4
3/8 inch
1/8
1/4
3/8
# Marprene TL
16
17
902.0032.PFQ
902.0064.PFQ
122 902.0096.PFQ
Pumpsil Neoprene
913.A032.PFQ
913.A064.PFQ
913.A096.PFQ
920.0032.PFQ
920.0064.PFQ
920.0096.PFQ
STA-PURE
Series PFL
965.0032.PFQ
965.0064.PFQ
965.0096.PFQ
# Bioprene TL
16
17
933.0032.PFT
933.0064.PFT
122 933.0096.PFT
Pumpsil
913.A032.PFT
913.A064.PFT
913.A096.PFT
STA-PURE
Series PCS
920.0032.PFT
920.0064.PFT
920.0096.PFT
STA-PURE
Series PFL
965.0032.PFT
965.0064.PFT
965.0096.PFT
2-4 bar (30-60 psi) pressure rated elements
Industrial mm
3.2
6.4
Sanitary mm
3.2
6.4
inch
1/8
1/4 inch
1/8
1/4
#
16
17
#
16
17
Marprene TM
902.P032.PFQ
902.P064.PFQ
Bioprene TM
933.P032.PFT
933.P064.PFT
STA-PURE Series
PFL
965.M032.PFQ
965.M064.PFQ
STA-PURE Series
PCS
960.M032.PFT
960.M064.PFT
74 m-530sn-gb-02
4-7 bar (60-100 psi) pressure rated elements
Industrial mm
3.2
Sanitary mm
3.2
inch
1/8 inch
1/8
#
16
#
16
Marprene TM
902.H032.PFQ
Bioprene TM
933.H032.PFT
STA-PURE Series
PCS
960.H032.PFQ
STA-PURE Series
PCS
960.H032.PFT
m-530sn-gb-02 75
25.3
Pumphead spares
1
2
4
5
6
7
3
520R/520R2
053.1011.100
053.1011.2L0
053.1011.EL0
053.1011.EM0
053.1011.EH0
MNA2050A (520R, 520R2)
MNA2045A (520R, 520R2)
520R
520REL/520REM/520REH
520R2
520REL
520REM
520REH
Pumphead guard complete with toolunlockable latch
Track assembly for cased pumps complete with spring-loaded tube clamps
MNA2043A (520R - 1.6mm wall tube)
MNA2001 (520R2 - 2.4mm wall tube)
MNA2138A (grey) (520REL)
MNA2139A (beige) (520REM)
MNA2140A (blue) (520REH)
MNA2006A (520R, 520R2)
MN2002M (520R, 520R2)
MN2131M (520RE)
MNA2005A (520R, 520R2)
MN2002M (520R, 520R2)
MN2034B
MN2005M
MNA2147A (520RE)
Rotor assembly complete with pumping rollers, follower rollers and tube guide rollers
Bottom (LH) tube clamp
Tube clamp location plug
Drain plug
Top (RH) tube clamp
Tube clamp location plug
Guard latch spring
Guard latch spring cartridge
Pumphead guard complete with seal and tool-lockable latch
76 m-530sn-gb-02
520R/520R2
MNA2144A (520RE)
8
9 MN2023T and MN2003T (520RE)
Parts may be ordered individually.
520REL/520REM/520REH
Track assembly for cased pumps
Drain port and nut m-530sn-gb-02 77
25.4
Pumping accessories
Accessory
520ANC
520ANX
520ANA
520AB
520AF
520AH
505LTC
505AS
520AL
505AFN
520AD
Description
Network cable, RS232, with 9-pin D-connectors
Network extension cable with 9-pin Dconnectors
Network adaptor, 25pin to 9-pin Dconnectors
Batch records cable with 9-pin D-connectors
Foot switch with 25-pin
D-connector
Hand switch with 25-pin
D-connector
Tube clamp set for 505L pumphead
Filling stand
Dispensing lance for use with 520AFN filling needles
Filling needle set
Filling needle 1.6mm
bore
Filling needle 3.2mm
bore
Filling needle 4.8mm
bore
Filling needle 6.4mm
bore
Filling needle 8.0mm
bore
Tube monitor with 25pin D-connector
Tube monitor, bare lead
Leak detector kit
Part code
059.3121.000
059.3122.000
059.3123.000
059.3125.000
059.3002.000
059.3022.000
059.4001.000
059.5001.000
059.5052.000
059.5101.000
059.5100.016
059.5100.032
059.5100.048
059.5100.064
059.5100.080
059.4501.520
059.450N.520
059.8121.000
Pump compatibility
530Du, 530Di
530Du, 530Di
530Di
530U, 530Du, 530Di
530U, 530Du, 530Di
530Di
All models
All models
All models
All models
All models
All models
All models
All models
530U, 530Du, 530Di
530UN, 530DuN,
530DiN
530U, 530Du, 530Di
78 m-530sn-gb-02
Accessory
520AD
Description
Leak detector kit
Part code
059.8131.000
Pump compatibility
530UN, 530DuN,
530DiN m-530sn-gb-02 79
26 Performance data
26.1
Performance curves
Flow rates of suction and discharge pressures for the pumphead at different drive speeds.
This data was produced pumping water at ambient temperature.
Marprene continuous tubing, 1.6mm wall, 200rpm, clockwise rotation
80 m-530sn-gb-02
Marprene continuous tubing, 1.6mm wall, 200rpm, counter-clockwise rotation
Marprene continuous tubing, 2.4mm wall, 200rpm, clockwise rotation
m-530sn-gb-02 81
Marprene continuous tubing, 2.4mm wall, 200rpm, counter-clockwise rotation
Marprene TL element, 0-2 bar (0-30psi), 200rpm, counter-clockwise rotation
82 m-530sn-gb-02
Sta-Pure element, 0-2 bar (0-30psi), 200rpm, counter-clockwise rotation
Marprene TM element, 2-4 bar (20-60psi), 200rpm, counter-clockwise rotation
m-530sn-gb-02 83
Sta-Pure element, 2-4 bar (30-60psi), 200rpm, counter-clockwise rotation
Marprene TH element, 4-7 bar(60-100psi), 200rpm, counter-clockwise rotation
84 m-530sn-gb-02
27 Trademarks
Watson-Marlow, Bioprene, Pumpsil, LoadSure and Marprene are trademarks of Watson-Marlow Limited.
Fluorel is a trademark of 3M.
STA-PURE Series PCS and STA-PURE Series PFL are trademarks of W.L.Gore and Associates.
m-530sn-gb-02 85
28 Disclaimers
The information contained in this document is believed to be correct but Watson-Marlow Fluid Technology
Group accepts no liability for any errors it contains and reserves the right to alter specifications without notice.
WARNING: This product is not designed for use in and should not be used for, patient- connected applications.
86 m-530sn-gb-02
29 Publication history
m-530sn-gb-02 530 S/SN pump
First published 12.15. Issue 2: 06.16.
m-530sn-gb-02 87
advertisement
Related manuals
advertisement