Watson Marlow Pumps 530 Bp, 530 Du, 530 S, 530 U Operating Manual

Add to My manuals
87 Pages

advertisement

Watson Marlow Pumps 530 Bp, 530 Du, 530 S, 530 U Operating Manual | Manualzz

530 S/SN Operating Manual

Contents

1 Declaration of conformity

2 Declaration of incorporation

3 When you unpack your pump

3.1 Unpacking your pump

3.2 Packaging disposal

3.3 Inspection

3.4 Components supplied

3.5 Storage

4 Information for returning pumps

5 Peristaltic pumps - an overview

6 Warranty

7 Safety notes

8 Pump specifications

8.1 Specification ratings

8.2 Weights

8.3 Pumphead options

9 Good pump installation practice

9.1 General recommendations

9.2 Dos and don'ts

10 Pump operation

10.1 Keypad Layout and Key IDs

10.2 Starting and stopping

10.3 Using up and down keys

10.4 Maximum speed

10.5 Change rotation direction

11 Connecting to a power supply

11.1 Conductor colour coding

12 Start-up check list

13 Switching the pump on for the first time

13.1 Selecting the display language

13.2 First-time start-up defaults

m-530sn-gb-02

18

19

20

21

17

18

18

18

15

15

16

17

13

13

13

14

22

22

23

9

10

7

8

6

6

6

6

6

6

4

5

1

2

14 Switching the pump on in subsequent power cycles

15 Mode menu

16 Manual

16.1 START

16.2 STOP

16.3 INCREASING AND DECREASING FLOW RATE

16.4 MAX FUNCTION (Manual mode only)

17 Flow calibration

17.1 Setting the flow calibration

18 MemoDose mode

18.1 To configure MemoDose

18.2 Set flow rate

18.3 Resume interrupted doses

18.4 Master dose

18.5 Manual dosing

19 Main menu

19.1 Security settings

19.2 General settings

19.3 Control settings

19.4 Help

20 Troubleshooting

20.1 Error codes

20.2 Technical support

21 Drive maintenance

22 Drive spares

23 Pumphead replacement

23.1 Pumphead replacement

24 Tube replacement

24.1 Continuous tubing

24.2 Tube elements

25 Ordering information

25.1 Pump part numbers

25.2 Tubing and element part numbers

25.3 Pumphead spares

25.4 Pumping accessories

26 Performance data

26.1 Performance curves

64

65

66

67

61

62

63

63

40

41

42

51

33

34

35

36

25

26

29

30

30

33

27

27

28

28

71

71

72

76

67

68

68

69

78

80

80

m-530sn-gb-02

27 Trademarks

28 Disclaimers

29 Publication history

85

86

87

m-530sn-gb-02 3

4

1 Declaration of conformity

Watson-Marlow Limited

Falmouth

Cornwall

TR11 4RU

England

1. 530 Cased pumps (Models: S, SN, U, UN, Du, DuN, Bp, BpN)

2. Manufacturer:

Watson Marlow Ltd

Bickland Water Road

Falmouth

EC Declaration of Conformity

TR11 4RU

UK

3. This declaration of conformity is issued under the sole responsibility of the manufacturer

4. All models and versions of the 530 series of cased peristaltic pump with all approved pump heads, tubing and accessories.

5. The object of the declaration described above is in conformity with the relevant Union harmonisation legislation:

Machinery Directive 2006/42/EC

EMC Directive 2014/30/EC (from 20 th

ROHS Directive 2011/65/EU

April 2016) and 2004/108/EC (until 19 th April 2016)

6. Harmonised standards used:

BS EN61010-1:2010 third edition Safety requirements for electrical equipment for measurement, control, and laboratory use Part 1: General requirements

EN61326-1:2013 Electrical equipment for measurement, control and laboratory use – EMC requirements Part 1: General requirements

BS EN 60529:1992+A2:2013 Degrees of protection provided by enclosures (IP code)

7. Intertek Testing and Certification Ltd, No: 3272281, performed compliance testing to BS EN

61010-1:2010, IEC 61010-1:2010, UL 61010-1:2010 and CAN/CSA C22.2 Bo 6101010-

1:2010 and issued certification of compliance to these standards.

Signed for and behalf of:

Watson Marlow Ltd

Falmouth, 6 th January 2016

Simon Nicholson, Managing Director, Watson-Marlow Limited m-530sn-gb-02

2 Declaration of incorporation

Watson-Marlow Ltd

Falmouth

Cornwall

TR11 4RU

England

Declaration of Incorporation

In accordance with the Machinery Directive 2006/42/EC that if this unit is to be installed into a machine or is to be assembled with other machines for installations, it shall not be put into service until the relevant machinery has been declared in conformity.

We hereby declare that:

Peristaltic Pump

Series: 530 cased pumps the following harmonised standards have been applied and fulfilled for health and safety requirements:

Safety of Machinery – EN ISO 12100

Safety of Machinery – Electrical Equipment of Machines BS EN 60204-1

Quality Management System – ISO 9001 and the technical documentation is compiled in accordance with Annex VII(B) of the Directive.

We undertake to transmit, in response to a reasoned request by the appropriate national authorities, relevant information on the partly completed machinery identified above. The method of transmission shall be by mail or email.

The pump head is incomplete and must not be put into service until the machinery into which it is to be incorporated has been declared in conformity with the provisions of the Directive.

Person authorised to compile the technical documents:

Andrew Green, Design & Engineering Director, Watson-Marlow Ltd.

Place and date of declaration: Watson-Marlow Ltd, 31.07.2015

Responsible person:

Simon Nicholson, Managing Director, Watson-Marlow Ltd m-530sn-gb-02 5

3 When you unpack your pump

3.1

Unpacking your pump

Unpack all parts carefully, retaining the packaging until you are sure all components are present and in good order. Check against the components supplied list, below.

3.2

Packaging disposal

Dispose of packaging materials safely, and in accordance with regulations in your area. The outer carton is made of corrugated cardboard and can be recycled.

3.3

Inspection

Check that all components are present. Inspect components for damage in transit. If anything is missing or damaged, contact your distributor immediately.

3.4

Components supplied

530 Components

530 pump drive unit, fitted with 520R2 or other pumphead if specified as a pump

The designated power cable (attached to the pump drive unit)

A 520N module providing pump ingress protection to IP66, NEMA 4X, if a 530SN, 530UN, 530DuN or 530BpN.

Note: the module is attached for transit, but must be removed to allow wiring up, voltage selection and fuse inspection and then re-affixed before the pump is operated.

3.5

Product safety information booklet incorporating quick start manual

Storage

This product has an extended shelf life. However, care should be taken after storage to ensure that all parts function correctly. Please observe the storage recommendations and use-by dates which apply to tubing you may wish to bring into service after storage.

6 m-530sn-gb-02

4 Information for returning pumps

In compliance with the UK Health and Safety at Work Act and the Control of Substances Hazardous to Health

Regulations, you are required to declare the substances which have been in contact with product(s) you return to Watson-Marlow or it subsidiaries or distributors. Failure to do so will cause delays. Please ensure that you email us this information and receive a RGA (Returned Goods Authorisation) before you despatch the product(s). A copy of the RGA form must be attached to the outside of the packaging containing the product(s).

Please complete a separate decontamination certificate for each product and attach it to the outside of the packaging containing the product(s). A copy of the appropriate decontamination certificate can be downloaded from the Watson-Marlow website at www.wmftg.com

You are responsible for cleaning and decontaminating the product(s) before return.

m-530sn-gb-02 7

5 Peristaltic pumps - an overview

Peristaltic pumps are the simplest possible pump, with no valves, seals or glands to clog or corrode. The fluid contacts only the bore of a tube, eliminating the risk of the pump contaminating the fluid, or the fluid contaminating the pump. Peristaltic pumps can operate dry without risk.

How they work

A compressible tube is squeezed between a roller and a track on an arc of a circle, creating a seal at the point of contact. As the roller advances along the tube, the seal also advances. After the roller has passed, the tube returns to its original shape, creating a partial vacuum which is filled by fluid drawn from the inlet port.

Before the roller reaches the end of the track, a second roller compresses the tube at the start of the track, isolating a packet of fluid between the compression points. As the first roller leaves the track, the second continues to advance, expelling the packet of fluid through the pump’s discharge port. At the same time, a new partial vacuum is created behind the second roller into which more fluid is drawn from the inlet port.

Backflow and siphoning do not occur, and the pump effectively seals the tube when it is inactive. No valves are needed.

The principle may be demonstrated by squeezing a soft tube between thumb and finger and sliding it along: fluid is expelled from one end of the tube while more is drawn in at the other.

Animal digestive tracts function in a similar way.

Suitable applications

Peristaltic pumping is ideal for most fluids, including viscous, shear-sensitive, corrosive and abrasive fluids, and those containing suspended solids. They are especially useful for pumping operations where hygiene is important.

Peristaltic pumps operate on the positive displacement principle. They are particularly suitable for metering, dosing and dispensing applications. Pumps are easy to install, simple to operate and inexpensive to maintain.

8 m-530sn-gb-02

6 Warranty

Watson-Marlow Ltd ("Watson- Marlow") warrants this product to be free from defects in materials and workmanship for five years from the date of shipment, under normal use and service.

Watson-Marlow's sole responsibility and the customer's exclusive remedy for any claim arising out of the purchase of any product from Watson-Marlow is, at Watson-Marlow's option: repair, replacement or credit, where applicable.

Unless otherwise agreed in writing, the foregoing warranty is limited to the country in which the product is sold.

No employee, agent or representative of Watson- Marlow has the authority to bind Watson-Marlow to any warranty other than the foregoing unless in writing and signed by a director of Watson-Marlow. Watson-

Marlow makes no warranty of the fitness of its products for a particular purpose.

In no event: i.

shall the cost of the customer's exclusive remedy exceed the purchase price of the product; ii.

shall Watson- Marlow be liable for any special, indirect, incidental, consequential, or exemplary damages, however arising, even if Watson- Marlow has been advised of the possibility of such damages.

Watson-Marlow shall not be liable for any loss, damage, or expense directly or indirectly related to or arising out of the use of its products, including damage or injury caused to other products, machinery, buildings, or property. Watson- Marlow shall not be liable for consequential damages, including without limitation, lost profits, loss of time, inconvenience, loss of product pumped, and loss of production.

This warranty does not obligate Watson-Marlow to bear any costs of removal, installation, transportation, or other charges which may arise in connection with a warranty claim.

Watson-Marlow shall not be responsible for shipping damage of returned items.

Conditions

o o o o

Products must be returned by pre- arrangement to Watson- Marlow, or a Watson- Marlow approved service centre.

All repairs or modifications must have been made by Watson-Marlow Ltd, or a Watson-Marlow approved service centre or with the express permission in writing of Watson-Marlow, signed by a manager or director of Watson-Marlow.

Any remote control or system connections must be made in accordance to Watson- Marlow recommendations.

All PROFIBUS systems must be installed or certified by a PROFIBUS approved installation engineer.

Exceptions

o o o o

Consumable items including tubing and pumping elements are excluded.

Pumphead rollers are excluded.

Repairs or service necessitated by normal wear and tear or by lack of reasonable and proper maintenance are excluded.

Products which, in the judgement of Watson- Marlow, have been abused, misused, or subject to malicious or accidental damage or neglect are excluded.

o o o o

Failure caused by electrical surge is excluded.

Failure caused by incorrect or sub-standard system wiring is excluded.

Damage by chemical attack is excluded.

Ancillaries such as leak detectors are excluded.

o o

Failure caused by UV light or direct sunlight.

Any attempt to disassemble a Watson-Marlow product will invalidate the product warranty.

Watson-Marlow reserves the right to amend these terms and conditions at any time.

m-530sn-gb-02 9

7 Safety notes

This safety information should be used in conjunction with the rest of this operating manual.

In the interests of safety, this pump and pumphead should only be used by competent, suitably trained personnel after they have read and understood the manual and considered any hazard involved. If the pump is used in a manner not specified by Watson- Marlow Ltd, the protection provided by the pump may be impaired. Any person who is involved in the installation or maintenance of this equipment should be fully competent to carry out the work. In the UK this person should also be familiar with the Health and Safety at

Work Act 1974.

This symbol, used on the pump and in the manual, means: Caution, refer to accompanying documents.

This symbol, used on the pump and in the manual, means: Do not allow fingers to contact moving parts.

This symbol, used on the pump and in the manual, means: Caution, hot surface.

This symbol, used on the pump and in the manual, means: Caution, risk of electric shock.

This symbol, used on the pump and in the manual, means: Personal

Protective Equipment (PPE) must be worn.

This symbol, used on the pump and in the manual, means: Recycle this product under the terms of the EU Waste Electrical and Electronic

Equipment (WEEE) Directive.

Fundamental work with regard to lifting, transportation, installation, starting-up, maintenance and repair should be performed by qualified personnel only. The unit must be isolated from mains power while work is being carried out. The motor must be secured against accidental start-up.

10 m-530sn-gb-02

Some pumps weigh more than 18kg (the exact weight depends on the model and pumphead - see on the pump). Lifting should be performed according to standard Health and Safety guidelines. Finger recesses are built into the sides of the lower shell for convenience in lifting; in addition, the pump can conveniently be lifted by grasping the pumphead and (where fitted) the 'N' module at the rear of the pump.

There is a user-replaceable type fuse in the fuse holder in the centre of the switch-plate at the back of the pump. In some countries, the mains power plug contains an additional replaceable fuse.

There are no user-serviceable fuses or parts inside this pump.

This pump must be used only for its intended purpose.

The pump must be accessible at all times to facilitate operation and maintenance. Access points must not be obstructed or blocked. Do not fit any devices to the drive unit other than those tested and approved by

Watson-Marlow. Doing so could lead to injury to persons or damage to property for which no liability can be accepted.

The pump's main plug is the disconnecting device (for isolating the motor drive from the mains supply in an emergency). Do not position the pump so that it is difficult to disconnect the mains plug.

If hazardous fluids are to be pumped, safety procedures specific to the particular fluid and application must be put in place to protect against injury to persons.

This product does not comply with the ATEX directive and must not be used in explosive atmospheres.

Ensure the chemicals to be pumped are compatible with the pumphead, lubricant (where applicable), tubing, pipework and fittings to be used with the pump. Please refer to the chemical compatibility guide which can be found at: www.wmftg.com/chemical .

If you need to use the pump with any other chemical please contact

Watson-Marlow to confirm compatibility.

m-530sn-gb-02 11

Pump will start as soon as power is applied if Auto restart is on and pump was running when power was disconnected.

There are moving parts inside the pumphead. Before opening the toolunlockable guard or tool-unlockable track, ensure that the following safety directions are followed:

1.

2.

3.

Ensure the pump is isolated from the mains power.

Ensure that there is no pressure in the pipeline.

4.

If a tube failure has occurred, ensure that any fluid in the pumphead has been allowed to drain to a suitable vessel, container or drain.

Ensure that appropriate Personal Protective Equipment (PPE) is worn.

Primary operator protection from rotating parts of the pump is provided by the pumphead safeguard. Note that safeguards differ, depending on the type of pumphead. See the pumphead section of the manual.

12 m-530sn-gb-02

8

8.1

Pump specifications

Specification ratings

Operating temperature

Storage temperature

Humidity (non-condensing)

Maximum altitude

Power consumption

Supply voltage

Maximum voltage fluctuation

Full load current

Fuse rating

Installation category

(overvoltage category)

Pollution degree

IP dB rating

5C to 40C (41F to 104F)

530: -40C to 70C (-40F to 158F)

80% up to 31C (88F) decreasing linearly to 50% at 40C (104F)

2000m (6560ft)

530: 135VA

Filtered 100-120V/200-240V 50/60Hz 1pH

+/-10% of nominal voltage. A well regulated electrical mains supply is required along with cable connections conforming to the best practice of noise immunity

530: <0.6A@ 230V; <1.25A @ 115V

530: Ceramic, 5x20mm, 2.5A, 250V AC, Time Delay

II

2

IP31 to BS EN 60529 if supplied with N module then IP66 to BS EN

60529. Equivalent to NEMA 4X to NEMA 250 *(indoor use - protect from prolonged UV exposure)

530: < 70dB (A) @ 1m

Control ratio

Maximum speed

8.2

Weights

530: 0.1-220rpm (2200:1)

530: 220rpm

530

IP31

IP66

Drive only

9.7kg

10.6kg

21lb 6oz

23lb 5oz

+ 520R, 520R2 + 520REL,520REM,

520REH

10.6kg

23lb 5oz 10.5kg

23lb 3oz

11.5kg

25lb 5oz 11.4kg

25lb 2oz

+ 505L

12.2kg

13.1kg

26lb 14oz

28lb 13oz

Weights warning

Some pumps weigh more than 18kg (the exact weight depends on the model and pumphead - see on the pump). Lifting should be performed according to standard Health and Safety guidelines. Finger recesses are built into the sides of the lower shell for convenience in lifting; in addition, the pump can conveniently be lifted by grasping the pumphead and (where fitted) the 'N' module at the rear of the pump.

m-530sn-gb-02 13

8.3

Pumphead options

530 pump range

520R, 501RL, 313, 314, 505L, 505BA, 505CA, 314MC, 318MC:

14 m-530sn-gb-02

9 Good pump installation practice

9.1

General recommendations

It is recommended that the pump is sited on a flat, horizontal, rigid surface, free from excessive vibration, to ensure the correct lubrication of the gearbox and correct pumphead operation. Allow a free flow of air around the pump to ensure that heat can be dissipated. Ensure that the ambient temperature around the pump does not exceed the recommended maximum operating temperature.

The STOP key on pumps supplied with a keypad will always stop the pump. However, it is recommended that a suitable local emergency stop device is fitted into the mains supply to the pump.

Do not stack the pumps more than the recommended maximum number. When the pumps are stacked, ensure that the ambient temperature around all the pumps in the stack does not exceed the recommended maximum operating temperature.

The pump may be set up so that the direction of rotor rotation is clockwise or counter-clockwise, whichever is convenient.

Please note, however, that for some pumpheads the tube life will be greater if the rotor rotates clockwise; and that performance against pressure will be maximised if the rotor rotates counter-clockwise. To achieve pressure in some pumpheads the pump must rotate counter-clockwise.

Peristaltic pumps are self- priming and self- sealing against backflow. No valves are required in inlet or discharge line, except those specified as below.

Users must fit a non-return valve between the pump and the discharge pipework to avoid the sudden release of pressurised fluid in the event of a pumphead or tube failure. This shall be fitted immediately after the discharge of the pump.

Valves in the process flow must be opened before the pump operates. Users are advised to fit a pressure relief device between the pump and any valve on the discharge side of the pump to protect against damage caused by accidental operation with the discharge valve closed.

m-530sn-gb-02 15

9.2

Dos and don'ts

l l l l l l l l l

Do not build a pump into a tight location without adequate airflow around the pump.

Do keep delivery and suction tubes as short and direct as possible - though ideally not shorter than one metre - and follow the straightest route. Use bends of large radius: at least four times the tubing diameter. Ensure that connecting pipework and fittings are suitably rated to handle the predicted pipeline pressure. Avoid pipe reducers and lengths of smaller bore tubing than the pumphead section, particularly in pipelines on the suction side. Any valves in the pipeline must not restrict the flow. Any valves in the flow line must be open when the pump is running.

Do ensure that on longer tube runs at least one metre of smooth bore, flexible tubing is connected to the inlet and discharge port of the pumphead to help to minimise impulse losses and pulsation in the pipeline.

This is especially important with viscous fluids and when connecting to rigid pipework.

Do use suction and delivery pipes equal to or larger than the tubing diameter bore. When pumping viscous fluids use pipe runs with a bore several times larger than the pump tube.

Do site the pump at or just below the level of the fluid to be pumped if possible. This will ensure flooded suction and maximum pumping efficiency.

Do run at slow speed when pumping viscous fluids. Flooded suction will enhance pumping performance, particularly for materials of a viscous nature.

Do recalibrate after changing tubing, fluid or any connecting pipework. It is also recommended that the pump is recalibrated periodically to maintain accuracy.

Do not pump any chemical not compatible with the tube or pumphead.

Do not run the pump with no tube or element fitted to the pumphead.

l l l l

Do not strap the control and mains cable together.

Do ensure if your product has an N module, that the module is fitted with the seals intact and properly located. Ensure that the holes for the cable glands are properly sealed to maintain the IP/NEMA rating.

If your pump is a PROFIBUS variant do ensure that it is installed in accordance with PROFIBUS installation guidelines.

Do not allow tight bends in any PROFIBUS signal cable.

Tube selection: The chemical compatibility guide published on the Watson Marlow website is for guidance. If in any doubt about the compatibility of a tube material and the duty fluid, request a Watson- Marlow tube sample card for immersion trials.

When using Marprene or Bioprene continuous tubing, do re-tension the tube after the first 30 minutes of running.

16 m-530sn-gb-02

10 Pump operation

10.1

Keypad Layout and Key IDs

HOME key

When the HOME key is pressed it will return the user to the last known operating mode. If modifying pump settings when the HOME key is pressed, it will disregard any setting changes and return you to the last known operating mode.

FUNCTION keys

FUNCTION keys, when pressed, will perform the function displayed on the screen directly above the relevant function key.

∧ and

∨ keys

These keys are used to change the programmable values within the pump. These keys are also used to move the selection bar up and down in the menus.

MODE key

To change modes or mode settings, press the MODE key. The MODE key can be pressed at any time to enter the mode menu. If modifying pump settings when the MODE key is pressed, it will disregard any setting changes and return you to the MODE menu.

m-530sn-gb-02 17

10.2

Starting and stopping

10.3

Using up and down keys

10.4

Maximum speed

10.5

Change rotation direction

18 m-530sn-gb-02

11 Connecting to a power supply

A well regulated electrical mains supply is required along with cable connections conforming to the best practice of noise immunity. It is not recommended to site these drives alongside 'dirty' electrical mains devices, such as 3- phase contactors and inductive heaters without special attention being paid to unacceptable mains-borne noise.

Set the voltage selector to 115V for 100-120V 50/60Hz supplies or

230V for 200-240V 50/60Hz supplies. Always check the voltage selector switch before connecting to the mains supply or the pump will be damaged.

Make suitable connection to an earthed single-phase mains electricity supply.

If the pump type is one with an 'N' module, the voltage selector is not visible while the module is in place. It is mounted in the switch plate at the rear of the pump, protected from water by the 'N' module. The module must be removed to allow access to the switch plate. Do not switch the pump on unless you have checked that it is set to suit your power supply by removing the module and inspecting the switch and then refitting the module.

1.

4.

7.

2.

5.

8.

3.

6.

9.

10.

We recommend using a commercially available supply voltage surge suppression where there is excessive electrical noise.

m-530sn-gb-02 19

Ensure that all power supply cables are adequately rated for the equipment.

The pump must be positioned so that the disconnection device is easily accessible when the equipment is in use.

11.1

Conductor colour coding

Conductor type

Line

Neutral

Ground

European colour

Brown

Blue

Green/Yellow

North American colour

Black

White

Green

20 m-530sn-gb-02

12 Start-up check list

Note : See also Tube loading.

o o o

Ensure that proper connections are achieved between the pump and suction and discharge piping.

Ensure proper connection has been made to a suitable power supply.

Ensure that the recommendations in the section are followed.

"Good pump installation practice" on page 15

m-530sn-gb-02 21

13 Switching the pump on for the first time

Power up the pump. The pump displays the start-up screen with the Watson-Marlow

Pumps logo for three seconds.

13.1

Selecting the display language

1. Use the and press

/

∨ keys to select your chosen language,

SELECT .

2. Your selected language will now be displayed on sceen. Choose CONFIRM to continue. All text will now appear in your chosen language.

3. Choose REJECT to return to the language choice screen. This then proceeds to the home screen.

22 m-530sn-gb-02

13.2

First-time start-up defaults

The pump is preset with operational parameters as shown in table below.

Parameter

Language

Default mode

Default manual speed

Pump status

Max speed

Direction

Pumphead

Tube size

Tube material

Flow calibration

Flow units

SG value

Keypad lock

Auto-restart

Analog signal type

Analog scaling type

Analog min current

Analog max current

Analog min flow rate/rpm

Analog max flow rate/rpm

Beeper

Security code

MemoDose flow rate

MemoDose volume

Pump number baud rate

Stop bits

Profibus node address

Remote start/stop input

Leak detector input

Input 4

Input 5

Output 1

Output 1 - status

530 default

Not set

Manual

220rpm

Stopped

220rpm

CW

520R2

9.6mm

Bioprene

15.12 ml/rev rpm

1

Disabled

OFF mA mA

5mA

19mA

0rpm

220rpm

ON

Not set

Mid flow rate of chosen pumphead

100ml

9600

2

126

High = stop

High = leak

Disabled

Disabled

Run/stop

High = run m-530sn-gb-02 23

Parameter

Output 2

Output 2 - status

Output 3

Output 3 - status

Output 4

Output 4 - status

Direction

High = CW

Auto/man

High = auto

General alarm

High = alarm

530 default

The pump is now ready to operate according to the defaults listed above.

Note: The display background colour changes according to running state as follows: l l

White background indicates pump stopped

Grey background indicates pump running l

Red background indicates error or alarm

All operating parameters may be changed by means of key-presses (see section

"Pump operation" on page 17

).

24 m-530sn-gb-02

14 Switching the pump on in subsequent power cycles

Subsequent power-up sequences will jump from the start-up screen to the home screen.

l l l

The pump runs a power-on test to confirm proper functioning of the memory and hardware. If a fault is found, an error code is displayed.

The pump displays the start-up screen with the Watson-Marlow Pumps logo for three seconds followed by the home screen

Start-up defaults are those in place when the pump was switched off last

Check that the pump is set to operate as you require it. The pump is now ready to operate.

All operating parameters may be changed by means of key- presses (see

"Pump operation" on page

17 ).

Power interruption

This pump has an auto restart feature which, when active, will restore the pump to the operating state it was in when power was lost.

Stop/start power cycles

Do not power up/power down the pump more than 20 times in an hour, whether manually or by means of the auto-restart facility. We recommend remote control where a high frequency of stop/start cycles is required.

Pump will start as soon as power is applied if Auto restart is on and pump was running when power was disconnected.

m-530sn-gb-02 25

15 Mode menu

Press MODE to display the Change mode menu.

Use the

∧ and

∨ keys to scroll through the available modes: o o o o

Manual (default)

Flow calibration

MemoDose

CANCEL

Use SELECT to choose mode. Use the right hand function key to alter mode settings.

26 m-530sn-gb-02

16 Manual

All settings and functions of the pump in manual mode are set and controlled by means of key- presses.

Immediately after the start- up display sequence detailed in:

"Switching the pump on in subsequent power cycles" on page 25

, the manual mode home screen will be displayed unless auto restart is enabled.

If enabled, the pump will return to the last known operating state when the power was lost. When the pump is running it displays an animated clockwise arrow. In normal operation, the direction of flow is into the bottom port of the pumphead and out of the top port.

If an exclamation mark (!) is displayed, it indicates that Auto restart is on (see

"Main menu" on page

41 ). If a padlock icon shows, it indicates that the keypad lock is on.

16.1

START

Starts the pump at the current flow displayed, and the display background changes to grey. If the pump is running, pressing this has no effect.

m-530sn-gb-02 27

16.2

STOP

Stops the pump. The display background changes to white. If the pump is not running pressing this has no effect.

16.3

INCREASING AND DECREASING FLOW RATE

Using the

∧ and

∨ keys will increase or decrease the flow rate.

Decreasing flow rate

o o o

A single key press will decrease flow rate by the least significant digit of the chosen flow rate unit.

Repeat key presses as required to achieve the desired flow rate.

Hold down the key for flow rate scrolling.

28 m-530sn-gb-02

Increasing flow rate

o o o

A single key press will increase flow rate by the least significant digit of the chosen flow rate unit.

Repeat key presses as required to achieve the desired flow rate.

Hold down the key for flow rate scrolling.

16.4

MAX FUNCTION (Manual mode only)

o o o Press and hold the MAX key to run at maximum flow.

Release the key to stop the pump.

The volume dispensed and time elapsed are displayed while the MAX key is pressed and held.

m-530sn-gb-02 29

17 Flow calibration

This pump displays flow rate in ml/min.

17.1

Setting the flow calibration

Using the

/

∨ keys, scroll to Flow calibration and press CALIBRATE .

Using the

/

∨ keys, enter the maximum flow rate limit and press ENTER .

30 m-530sn-gb-02

Press START to begin pumping a volume of fluid for calibration.

Press STOP to stop pumping fluid for the calibration.

m-530sn-gb-02 31

Use the

/

∨ keys to enter the actual volume of fluid pumped.

To accept the new calibration press or MODE to abort.

ACCEPT or RE-CALIBRATE to repeat the procedure. Press HOME

The pump is now calibrated.

32 m-530sn-gb-02

18 MemoDose mode

Each time the pump is started by pressing START , it records the number of pumphead revolutions which occur until STOP is pressed. The number of revolutions is proportional to the volume of fluid which has been dispensed: the dose. MemoDose mode allows the user to repeat-dose a precise volume of fluid. This can be achieved by dispensing a master dose, or entering a dose volume manually using the keypad. MemoDose can repeat this dose exactly or proportionately.

18.1

To configure MemoDose

o o

Select MODE

Using the

/

∨ keys, scroll to MemoDose and press SETTINGS

Note: the pump must be stopped to enter MemoDose Settings.

m-530sn-gb-02 33

18.2

Set flow rate

Using the

/

∨ keys, scroll to Flow rate and press SELECT .

Using the

/

∨ keys, enter the dose flow rate and press SELECT .

34 m-530sn-gb-02

18.3

Resume interrupted doses

MemoDose Mode gives you the opportunity to resume interrupted doses following a power cycle (Note:

Auto-restart must be on for this to function). Alternatively interrupted doses can be discarded and a new dose started when power is resumed.

From the MemoDose settings screen, use the press

/

∨ keys to scroll to Resume Interrupted Dose and

ENABLE . The red cross will become a green tick to indicate that Resume interrupted dose is enabled.

If this function is enabled the ENABLE key becomes DISABLE . If this is pressed interrupted doses will be discarded after a power cycle.

If resume interrupted doses has been set the following warning screen is displayed, press save this setting.

CONFIRM to m-530sn-gb-02 35

If resume interrupted doses has NOT been set the following warning screen is displayed, press to save this setting.

CONFIRM

18.4

Master dose

Using the

/

∨ keys, scroll to Master dose and press SELECT .

36 m-530sn-gb-02

The following screen is displayed; press master dose.

MANUAL to enter a dose via the keypad or DOSE to dispense a

Dispensing a master dose

Press START to begin dispensing the Master Dose.

m-530sn-gb-02 37

Press STOP to finish dispensing the Master Dose.

Manual dose entry

Press MANUAL on the master dose screen. Using the

/

∨ keys enter the required dose volume and press

SELECT to record the master dose or CANCEL to return to MemoDose Settings.

38 m-530sn-gb-02

Save dose volume

Press SAVE to record the Master Dose or CANCEL to return to MemoDose settings.

When MemoDose SETTINGS are complete the following screen is displayed; press

MemoDose Mode or BACK to return to MemoDose settings.

MEMODOSE to start in m-530sn-gb-02 39

18.5

Manual dosing

From the MemoDose Home screen, press START to deliver a dose. The screen displays the dose flow rate and the dose remaining counting down from 100% to 0%.

If the delivered dose differs from the volume required, the percentage may be adjusted within the limits 1% to 999% of the master dose. Use the

/

∨ keys to alter the percentage. The new dose size is displayed as a percentage on the home screen.

If STOP is pressed during dosing, the pump stops. Pressing START will resume, or discard, the interrupted dose depending on the settings in

"MemoDose mode" on page 33 .

40 m-530sn-gb-02

19 Main menu

To access the main menu press the MENU button from one of the HOME screens or INFO screens.

This will display the main menu as shown below. Use the available options.

/

∨ keys to move the selection bar between the

Press SELECT to choose an option.

Press EXIT to return to the screen from where the MENU was called.

m-530sn-gb-02 41

19.1

Security settings

Security settings can be changed by selecting SECURITY SETTINGS from the Main menu.

Auto keypad lock

Press ENABLE / DISABLE to switch on/off the Auto keypad lock. When active the keypad will 'lock' after

20 seconds of inactivity.

Once locked it will display the screen below when any key is pressed. To unlock the keypad press the two

UNLOCK keys together.

42 m-530sn-gb-02

The padlock icon will appear on the operating mode home screen to show that keypad lock is activated.

Note that the STOP key will always work whether the keypad is locked or not.

PIN protection

Using the

/

∨ keys select PIN protection from the SECURITY SETTINGS menu and press

ENABLE/DISABLE to switch on/off the PIN protection. If PIN protection has been enabled, a Master level

PIN will be required to disable PIN lock.

Setting Master PIN

Setting the Master PIN protects all functionality. The Master is able to selectively enable functionality for two additional operators. These are defined as User 1 and User 2. They will be able to access this functionality by entering a PIN code assigned to them by the Master user. To set the Master PIN, scroll to Master level and press ENABLE .

m-530sn-gb-02 43

To define a four digit Master PIN, use the digit press the

/

∨ keys to select each digit from 0-9. Once you have the required

NEXT DIGIT key. After selecting the fourth digit press ENTER .

Now press CONFIRM to PIN entry.

to check that the number entered is the PIN you require. Press CHANGE to return

44 m-530sn-gb-02

The following screen will be displayed to indicate that the Master PIN has been applied to access all functionality. Press NEXT to selectively enable functionality access for User 1 and User 2.

Configure User 1 security settings

The PIN PROTECTION level screen will be displayed with User 1 highlighted, press

User 1 security settings or scroll to configure an alternative User.

ENABLE to configure m-530sn-gb-02 45

ENABLE user 1 security settings displays the PIN entry screen for User 1. To define a four digit User 1 PIN, use the

/

∨ keys to select each digit from 0-9. Once you have the required digit press the NEXT DIGIT key. After selecting the fourth digit press ENTER .

Now press CONFIRM to PIN entry.

to verify that the number entered is the PIN you require. Press CHANGE to return

46 m-530sn-gb-02

To define the allowed functionality, use the

/

∨ keys to select the functionality and press ENABLE . User 1

PIN will allow access to only the enabled functionality, to disable functionality, highlight the enabled functionality and press DISABLE . When all the required functionality has been enabled, press FINISH .

Configure User 2 security settings

The PIN PROTECTION level screen will be displayed with User 2 highlighted, press

User 2 security settings or scroll to configure an alternative User.

ENABLE to configure m-530sn-gb-02 47

ENABLE user 2 security settings displays the PIN entry screen for User 2. To define a four digit User 2 PIN, use the

/

∨ keys to select each digit from 0-9. Once you have the required digit press the NEXT DIGIT key. After selecting the fourth digit press ENTER .

To define the allowed functionality, use the

/

∨ keys to select the functionality and press ENABLE . User 2

PIN will allow access to only the enabled functionality, to disable functionality, highlight the enabled functionality and press DISABLE . When all the required functionality has been enabled, press FINISH .

48 m-530sn-gb-02

Note : Once Security Settings for User 1 and User 2 have been set by the Master, only the Master PIN will allow access to Security Settings.

The HOME screen will be displayed. A PIN is now required to access all functionality. The Master PIN accesses all pump functionality and the User 1 and User 2 PINs access only the defined functionality. To enter the PIN, use the

/

∨ keys to select each digit from 0-9. Once you have the required digit press the

NEXT DIGIT key. After selecting the fourth digit press ENTER .

If an incorrect PIN has been entered the following screen will be displayed. NOTE: this screen will also display if the PIN entered does not allow access to that functionality.

If a PIN number is entered that is already in use, the following screen will be displayed, press input an alternative PIN or EXIT to abort

CHANGE to m-530sn-gb-02 49

If the PIN entered does not allow access to the functionality the following screen will be displayed.

Keypad beep

From SECURITY settings scroll to Keypad beep using the beep at every key press.

/

∨ keys and select ENABLE . The pump will now

50 m-530sn-gb-02

19.2

General settings

To view the general settings menu, select GENERAL SETTINGS from the main menu.

Auto restart

This pump offers an auto restart feature. If active on power loss, it restores the pump when power returns to the operating state it was in when power was lost.

For example, if the pump was running in analog mode prior to power loss, it would return to the same operating mode and continue to run at a proportional speed to the analog input.

Press ENABLE / DISABLE to turn the auto restart feature on/off.

Do not use auto restart for more than 20 mains power starts per hour.

We recommend remote control where a high number of starts is required.

Pump will start automatically if start conditions are met.

m-530sn-gb-02 51

The !

icon is displayed on the home screens to indicate that the auto restart feature is active.

Flow units

The current chosen flow unit is displayed on the right hand side of the screen. To change flow units move the selection bar over the flow unit menu entry and press SELECT .

Use the

/

∨ keys to move the selection bar over the required flow unit, then press SELECT . All flow rates displayed on screens will now be in the chosen units.

52 m-530sn-gb-02

If a mass flow unit is selected, the specific gravity of the fluid must be entered. The following screen is displayed.

Use the

/

∨ keys to enter the value of the specific gravity, and press SELECT .

Pump label

The pump label is a user defined 20 digit alphanumeric label which is displayed in the header bar of the home screen. To define or edit the pump label, move the selection bar over the Pump label menu entry and press

SELECT . If a pump label has been previously defined, this will be displayed on screen to allow editing, otherwise it will display the default label “WATSON-MARLOW".

m-530sn-gb-02 53

Use the

/

∨ keys to scroll through the available characters for each digit. The available characters are 0- 9,

A-Z and SPACE.

Press NEXT to move onto the next character, or PREVIOUS to move back to the previous character.

Press FINISH to save the entry and return to the general settings menu.

54 m-530sn-gb-02

Pumphead type

Select GENERAL SETTINGS from the main menu.

Use the

/

∨ keys to move the selection bar over Pumphead type and press SELECT . The following screen will be displayed.

Use the

/

∨ keys to move the selection bar over Pumphead and press SELECT .

m-530sn-gb-02 55

Use the

/

∨ keys to move the selection bar over the required pumphead type and press SELECT .

Tube size and tube material

Select Tube size from GENERAL SETTINGS, then use the

/

∨ keys to move the selection bar over Bore size and press SELECT .

56 m-530sn-gb-02

Use the

/

∨ keys to move the selection bar over the tube size to be used and press SELECT .

If a LoadSure element has been selected then the tube size is displayed as pressure and bore.

m-530sn-gb-02 57

This screen also allows you to select the tube material used.

Use the

/

∨ keys to move the selection bar over Tube material and press SELECT .

Use the

/

∨ keys to move the selection bar over the tube material to be used and press SELECT .

58 m-530sn-gb-02

The PUMPHEAD MODEL screen allows the tube Lot Number to be recorded for future reference. Use the

/

∨ keys to move the selection bar over Tube lot number and press SELECT .

Use the

/

∨ keys to scroll through the available characters for each digit. The available characters are 0- 9,

A-Z, and SPACE.

Press NEXT to move onto the next character, or PREVIOUS to move back to the last character.

Press FINISH to save the entry and return to the general settings menu.

Restore defaults

To restore the factory default settings select Restore defaults from the general settings menu.

There are two confirmation screens to ensure that this function is not carried out in error.

Press CONFIRM followed by RE-CONFIRM to restore the defaults.

m-530sn-gb-02 59

Language

Select language from the general settings menu to choose an alternative display language for the pump. The pump must be stopped before changing the language.

Use the

/

∨ keys to move the selection bar to your required language. Press SELECT to confirm.

Your selected language will now be displayed on screen. Press CONFIRM to continue, all displayed text will now appear in your chosen language.

Press REJECT to return to the language choice screen.

MODE menu

Selecting MODE menu from the main menu will navigate you to access the sub- menu shown below. This is the same as pressing the MODE key. Please see

"Mode menu" on page 26

for further details.

60 m-530sn-gb-02

19.3

Control settings

Select CONTROL SETTINGS from the main menu to access the sub menu shown below. Use the

/

∨ keys to move the selection bar. Press SELECT to choose the required function.

Speed limit

The maximum speed the pump is capable of running at is 220rpm.

Select Speed limit from the control settings menu to define a lower maximum speed limit for the pump.

This speed limit will be applied to all operating modes.

Use the

/

∨ keys to adjust the value and press SAVE to set.

m-530sn-gb-02 61

Reset run hours

Select Reset run hours from the control settings menu.

Select RESET to zero the run hours counter. The run hours counter can be viewed by pressing INFO from your home screen. The following screen will be displayed. Press RESET to reset the run hours or CANCEL to return to the CONTROL SETTINGS menu.

19.4

Help

Select Help from the main menu to access the help screens.

62 m-530sn-gb-02

20 Troubleshooting

If the pump display remains blank when the pump is switched on, make the following checks:

Check that the mains power is available to the pump.

Check the fuse in the wall plug if one is present.

Check the position of the voltage selector switch.

Check the mains power switch at the rear of the pump.

Check the fuse in the fuse holder in the centre of the switch plate at the rear of the pump.

If the pump runs but there is little or no flow, make the following checks:

Check that fluid is supplied to the pump.

Check for any kinks or blockages in the lines.

Check that any valves in the line are open.

Check that the tube and rotor are in the pumphead.

Check that the tube is not split or burst.

Check that the correct wall-thickness tube is being used.

Check the direction of rotation.

Check that the rotor is not slipping on the drive shaft.

If the pump turns on, but will not run:

Check the remote stop function and configuration.

Check the mode you are in, are you in analog mode.

Try to operate and run the pump in manual mode.

20.1

Error codes

If an internal error occurs, an error screen with a red background is displayed. Note: Signal out of range, over signal and leak detected error screens report the nature of an external condition. They do not flash.

Error code Error condition

Er 0

Er 1

Er 2

Er 3

Er 4

Er 9

FRAM write error

FRAM corruption

FLASH write error during drive update

FLASH corruption

FRAM shadow error

Motor stalled

Suggested action

Attempt to reset by switching power OFF/ON. Or seek support.

Attempt to reset by switching power OFF/ON. Or seek support.

Attempt to reset by switching power OFF/ON. Or seek support.

Attempt to reset by switching power OFF/ON. Or seek support.

Attempt to reset by switching power OFF/ON. Or seek support.

Stop pump immediately. Check pumphead and tube.

Power OFF/ON may reset. Or seek support.

m-530sn-gb-02 63

Er15

Er16

Er17

Er19

Er20

Er21

Error code

Er10

Error condition

Tacho fault

Er14 Speed error

Er30

Over current

Over voltage

Under voltage

Over temperature

Signal out of range

Over signal

Over power

Err50 Communication error

20.2

Technical support

Watson-Marlow Fluid Technology Group

Falmouth, Cornwall

TR11 4RU

UK

Telephone: +44 (0) 1326 370370

Fax: +44 (0) 1326 376009

Email: [email protected]

www.wmftg.com

Suggested action

Stop pump immediately.

Power OFF/ON may reset. Or seek support.

Stop pump immediately.

Power OFF/ON may reset. Or seek support.

Stop pump immediately.

Power OFF/ON may reset. Or seek support.

Stop pump immediately. Check supply.

Power OFF/ON may reset.

Stop pump immediately. Check supply.

Power OFF/ON may reset.

Stop pump immediately. Turn OFF. Seek support.

Check analog control signal range. Trim signal as required. Or seek support.

Reduce the analog control signal.

Turn OFF. Check power supply. Check pumphead and tubing. Wait 30 minutes. Power ON may reset.

Or seek support.

Attempt to reset by switching power OFF/ON. Or seek support.

64 m-530sn-gb-02

21 Drive maintenance

There are no user serviceable parts inside the pump. The unit should be returned to Watson- Marlow for service.

m-530sn-gb-02 65

22 Drive spares

Description

Replaceable main fuse, type T2, 2.5A H 250V 20mm

Foot

Module seal

Module PCB(SCADA)

Module PCB (NEMA)

Module switch cover

Glands (STD)

Glands (EMC)

Blanking plugs

Sealing washer for blanking plug and gland

Snap-fit vent

Part No.

FS0064

MN2507M

MN2516B

MNA2066A

MNA2069A

MN2505M

GR0056

GR0075

GR0057

GR0058

MN2513B

66 m-530sn-gb-02

23 Pumphead replacement

Always isolate the pump from the mains power supply before opening any guard or track, or performing any positioning, removal or maintenance operation.

23.1

Pumphead replacement

520R

1.

2.

3.

4.

5.

6.

m-530sn-gb-02 67

1.

4.

24 Tube replacement

Always isolate the pump from the mains power supply before opening any guard or track, or performing any positioning, removal or maintenance operation.

24.1

Continuous tubing

520R and 520R2

2.

5.

3.

6.

7.

505L

8.

1.

2.

4.

5.

≤8.0mm=145mm ,

9.6mm=150mm

68

9.

3.

6.

m-530sn-gb-02

24.2

Tube elements

Grey

>2 bar (30psi)

Beige

>4 bar (60psi)

Blue

>7 bar (100psi)

520REL, 520REM and 520REH

1.

2.

4.

Sanitary connectors

5.

1.

Industrial connectors

2.

1.

2.

m-530sn-gb-02

3.

3.

3.

69

505L

1.

2.

3.

4.

General guide to cleaning with solvents

Chemical

Aliphatic hydrocarbons

Aromatic hydrocarbons

Ketone solvents

Halogenated/chlorinated solvents

Alcohols, general

Glycols

Estersolvents

Ether solvents

Cleaning precautions

Remove guard. Minimize rotor cap and clutch boot exposure to less than one minute (risk of attack).

Remove guard. Minimize rotor cap and clutch boot exposure to less than one minute (risk of attack).

Remove guard. Minimize rotor cap and clutch boot exposure to less than one minute (risk of attack).

Not recommended: possible risk to polycarbonate tube clamp adjusters and polypropylene tube clamp locators.

No precaution necessary.

Minimize rotor cap and clutch boot exposure to less than one minute (risk of attack).

Remove guard. Minimize rotor cap and tube clamp location cap exposure to less than one minute (risk of attack).

Not recommended: possible risk to polycarbonate tube clamp adjusters and polypropylene tube clamp locators.

70 m-530sn-gb-02

25 Ordering information

25.1

Pump part numbers

530 Part Code Configurator

0 5 ● 9 1 ● ● ● ● ●

Configuration

0: Pump

6: Drive only

Model

3: S

4: U

5: Du

6: Bp

Ingress

Protection*

1: IP31

N: IP66

*IP66 pumps are supplied with no plug or an

American mains plug.

Pumphead

EL: 520REL

EM: 520REM

EH: 520REH

2L: 520R2

10: 520R

5L: 505L

Plug options*

0: no plug

U: UK mains plug

E: EU mains plug

A: American mains plug

K: Australia mains plug

R: Argentina mains plug

C: Swiss mains plug

D

:

India/South

Africa mains plug

m-530sn-gb-02 71

25.2

Tubing and element part numbers

1.6mm wall tubing for 520R pumpheads

mm inch # Marprene Bioprene

4.8

6.4

8.0

mm

0.8

0.5

0.8

1.6

3.2

1.6

3.2

4.8

6.4

8.0

mm

0.5

0.8

1.6

3.2

4.8

6.4

8.0

1/50

1/32

1/16

1/8

3/16

1/4

5/16 inch

1/32

1/16

1/8

3/16

1/4

5/16 inch

1/50

1/32

1/16

1/8

3/16

1/4

5/16

112

13

14

16

25

17

18

#

13

14

16

25

17

18

#

112

13

14

16

25

17

18

902.0005.016

902.0008.016

902.0016.016

902.0032.016

902.0048.016

902.0064.016

902.0080.016

PVC

950.0016.016

950.0032.016

950.0048.016

950.0064.016

950.0080.016

Pumpsil

913.A005.016

913.A008.016

913.A016.016

913.A032.016

913.A048.016

913.A064.016

913.A080.016

933.0005.016

933.0008.016

933.0016.016

933.0032.016

933.0048.016

933.0064.016

933.0080.016

Fluorel

970.0016.016

970.0032.016

970.0048.016

970.0064.016

970.0080.016

STA-PURE

Series PCS

960.0016.016

960.0032.016

960.0048.016

960.0064.016

960.0080.016

Note: 1.6mm wall STA-PURE Series PFL and STA-PURE Series PCS tubing are supplied in 305mm lengths.

STA-PURE

Series PFL

965.0016.016

965.0032.016

965.0048.016

965.0064.016

965.0080.016

Neoprene

920.0008.016

920.0016.016

920.0032.016

920.0048.016

920.0064.016

920.0080.016

72 m-530sn-gb-02

2.4mm wall tubing for 520R2 pumpheads

mm

0.5

0.8

1.6

3.2

4.8

6.4

8.0

9.6

mm

0.8

1.6

3.2

4.8

6.4

8.0

inch

1/50

1/32

1/16

1/8

3/16

1/4

5/16

3/8 inch

1/32

1/16

1/8

3/16

1/4

5/16

#

119

120

15

24

121

122

#

119

120

15

24

121

Marprene

902.0016.024

902.0032.024

902.0048.024

902.0064.024

902.0080.024

902.0096.024

STA-PURE

Series PFL

965.0016.024

965.0032.024

965.0048.024

965.0064.024

965.0080.024

Bioprene

933.0016.024

933.0032.024

933.0048.024

933.0064.024

933.0080.024

933.0096.024

STA-PURE

Series PCS

960.0016.024

960.0032.024

960.0048.024

960.0064.024

960.0080.024

Pumpsil

913.A005.024

913.A008.024

913.A016.024

913.A032.024

913.A048.024

913.A064.024

913.A080.024

913.A096.024

Note: 2.4mm wall STA-PURE Series PFL and STA-PURE Series PCS tubing are supplied in 355mm lengths.

m-530sn-gb-02 73

2.4mm wall elements for 520RE pumpheads

0-2 bar (0-30 psi) pressure rated elements

Industrial mm

3.2

6.4

9.6

Sanitary mm

3.2

6.4

9.6

inch

1/8

1/4

3/8 inch

1/8

1/4

3/8

# Marprene TL

16

17

902.0032.PFQ

902.0064.PFQ

122 902.0096.PFQ

Pumpsil Neoprene

913.A032.PFQ

913.A064.PFQ

913.A096.PFQ

920.0032.PFQ

920.0064.PFQ

920.0096.PFQ

STA-PURE

Series PFL

965.0032.PFQ

965.0064.PFQ

965.0096.PFQ

# Bioprene TL

16

17

933.0032.PFT

933.0064.PFT

122 933.0096.PFT

Pumpsil

913.A032.PFT

913.A064.PFT

913.A096.PFT

STA-PURE

Series PCS

920.0032.PFT

920.0064.PFT

920.0096.PFT

STA-PURE

Series PFL

965.0032.PFT

965.0064.PFT

965.0096.PFT

2-4 bar (30-60 psi) pressure rated elements

Industrial mm

3.2

6.4

Sanitary mm

3.2

6.4

inch

1/8

1/4 inch

1/8

1/4

#

16

17

#

16

17

Marprene TM

902.P032.PFQ

902.P064.PFQ

Bioprene TM

933.P032.PFT

933.P064.PFT

STA-PURE Series

PFL

965.M032.PFQ

965.M064.PFQ

STA-PURE Series

PCS

960.M032.PFT

960.M064.PFT

74 m-530sn-gb-02

4-7 bar (60-100 psi) pressure rated elements

Industrial mm

3.2

Sanitary mm

3.2

inch

1/8 inch

1/8

#

16

#

16

Marprene TM

902.H032.PFQ

Bioprene TM

933.H032.PFT

STA-PURE Series

PCS

960.H032.PFQ

STA-PURE Series

PCS

960.H032.PFT

m-530sn-gb-02 75

25.3

Pumphead spares

1

2

4

5

6

7

3

520R/520R2

053.1011.100

053.1011.2L0

053.1011.EL0

053.1011.EM0

053.1011.EH0

MNA2050A (520R, 520R2)

MNA2045A (520R, 520R2)

520R

520REL/520REM/520REH

520R2

520REL

520REM

520REH

Pumphead guard complete with toolunlockable latch

Track assembly for cased pumps complete with spring-loaded tube clamps

MNA2043A (520R - 1.6mm wall tube)

MNA2001 (520R2 - 2.4mm wall tube)

MNA2138A (grey) (520REL)

MNA2139A (beige) (520REM)

MNA2140A (blue) (520REH)

MNA2006A (520R, 520R2)

MN2002M (520R, 520R2)

MN2131M (520RE)

MNA2005A (520R, 520R2)

MN2002M (520R, 520R2)

MN2034B

MN2005M

MNA2147A (520RE)

Rotor assembly complete with pumping rollers, follower rollers and tube guide rollers

Bottom (LH) tube clamp

Tube clamp location plug

Drain plug

Top (RH) tube clamp

Tube clamp location plug

Guard latch spring

Guard latch spring cartridge

Pumphead guard complete with seal and tool-lockable latch

76 m-530sn-gb-02

520R/520R2

MNA2144A (520RE)

8

9 MN2023T and MN2003T (520RE)

Parts may be ordered individually.

520REL/520REM/520REH

Track assembly for cased pumps

Drain port and nut m-530sn-gb-02 77

25.4

Pumping accessories

Accessory

520ANC

520ANX

520ANA

520AB

520AF

520AH

505LTC

505AS

520AL

505AFN

520AD

Description

Network cable, RS232, with 9-pin D-connectors

Network extension cable with 9-pin Dconnectors

Network adaptor, 25pin to 9-pin Dconnectors

Batch records cable with 9-pin D-connectors

Foot switch with 25-pin

D-connector

Hand switch with 25-pin

D-connector

Tube clamp set for 505L pumphead

Filling stand

Dispensing lance for use with 520AFN filling needles

Filling needle set

Filling needle 1.6mm

bore

Filling needle 3.2mm

bore

Filling needle 4.8mm

bore

Filling needle 6.4mm

bore

Filling needle 8.0mm

bore

Tube monitor with 25pin D-connector

Tube monitor, bare lead

Leak detector kit

Part code

059.3121.000

059.3122.000

059.3123.000

059.3125.000

059.3002.000

059.3022.000

059.4001.000

059.5001.000

059.5052.000

059.5101.000

059.5100.016

059.5100.032

059.5100.048

059.5100.064

059.5100.080

059.4501.520

059.450N.520

059.8121.000

Pump compatibility

530Du, 530Di

530Du, 530Di

530Di

530U, 530Du, 530Di

530U, 530Du, 530Di

530Di

All models

All models

All models

All models

All models

All models

All models

All models

530U, 530Du, 530Di

530UN, 530DuN,

530DiN

530U, 530Du, 530Di

78 m-530sn-gb-02

Accessory

520AD

Description

Leak detector kit

Part code

059.8131.000

Pump compatibility

530UN, 530DuN,

530DiN m-530sn-gb-02 79

26 Performance data

26.1

Performance curves

Flow rates of suction and discharge pressures for the pumphead at different drive speeds.

This data was produced pumping water at ambient temperature.

Marprene continuous tubing, 1.6mm wall, 200rpm, clockwise rotation

80 m-530sn-gb-02

Marprene continuous tubing, 1.6mm wall, 200rpm, counter-clockwise rotation

Marprene continuous tubing, 2.4mm wall, 200rpm, clockwise rotation

m-530sn-gb-02 81

Marprene continuous tubing, 2.4mm wall, 200rpm, counter-clockwise rotation

Marprene TL element, 0-2 bar (0-30psi), 200rpm, counter-clockwise rotation

82 m-530sn-gb-02

Sta-Pure element, 0-2 bar (0-30psi), 200rpm, counter-clockwise rotation

Marprene TM element, 2-4 bar (20-60psi), 200rpm, counter-clockwise rotation

m-530sn-gb-02 83

Sta-Pure element, 2-4 bar (30-60psi), 200rpm, counter-clockwise rotation

Marprene TH element, 4-7 bar(60-100psi), 200rpm, counter-clockwise rotation

84 m-530sn-gb-02

27 Trademarks

Watson-Marlow, Bioprene, Pumpsil, LoadSure and Marprene are trademarks of Watson-Marlow Limited.

Fluorel is a trademark of 3M.

STA-PURE Series PCS and STA-PURE Series PFL are trademarks of W.L.Gore and Associates.

m-530sn-gb-02 85

28 Disclaimers

The information contained in this document is believed to be correct but Watson-Marlow Fluid Technology

Group accepts no liability for any errors it contains and reserves the right to alter specifications without notice.

WARNING: This product is not designed for use in and should not be used for, patient- connected applications.

86 m-530sn-gb-02

29 Publication history

m-530sn-gb-02 530 S/SN pump

First published 12.15. Issue 2: 06.16.

m-530sn-gb-02 87

advertisement

Related manuals

advertisement