Roche cobas p 512 Below you will find brief product information for Roche cobas p 512. For detailed information on the product, functions, and usage instrcutions, please refer to the operator's manual.
Roche cobas p 512 can be used for:
- Pre-analytical sample preparation
- Automatic sorting of sample containers
- Aliquoting into primary and secondary tubes
- Capping of sample containers
- Automatic loading of secondary sample containers onto the analyzer
- Automatic unloading of primary sample containers
In addition, the following features of Roche cobas p 512 are worth mentioning:
- Intuitive user interface and touchscreen
advertisement
Addendum 1
For cobas p 512 pre-analytical system (47x) operator’s manual 1.6
24-position centrifuge rack
Addendum 1
24-position centrifuge rack
Publication information
Publication version
1.0
Revision date
Jan 2018
Change description
First release
Revisions
This chapter contains additional information for the cobas p 512 pre-analytical system
(47x) operator’s manual and was added due to additional information about the use of the
24-position centrifuge rack (RD ID 06623611001).
2
Addendum 1
24-position centrifuge rack
Table of contents
3
Addendum 1
24-position centrifuge rack
Affected system
System
Operator’s manual publication version *
Applicable software version
Change description cobas p 512
(SN 47xxxxxx)
1.6 1.0.0 and higher Information about the loading of the rack and the labelling of the tubes
* For US customers, this addendum is applicable to the currently approved version of the aforementioned operator’s manual found on the US Roche Diagnostics website.
Table 1: Affected system
Roche recommends that you familiarize yourself with the new and/or revised content provided in this addendum.
If you print these pages, Roche recommends printing them single sided. In this way, you can easily insert the new and/or revised content in its appropriate locations in the existing document.
Additional information
This information specifically refers to the usage of 24-position centrifuge racks.
Figure 1: 24-position centrifuge rack (RD ID 06623611001)
4
Addendum 1
24-position centrifuge rack
Loading of the 24-position centrifuge rack
This information specifically refers to the usage of 24-position centrifuge racks and is an addition to section 3.3.9 in the operator’s manual.
NOTICE
Damage to the system due to displaced spring insets
The gripper in the sorting area of the subsequent system may crash into the tube in cases where the spring insets of a 24-position centrifuge rack are displaced.
•
Do not use a 24-position centrifuge rack with displaced spring insets.
•
Avoid empty rack positions during centrifugation.
When a Hettich stand-alone centrifuge and 24-position centrifuge racks
(RD ID 06623611001) are used, make sure that:
•
24-position centrifuge racks are fully loaded for centrifuging.
•
Balance tubes are used to fill empty rack positions.
Placement of barcode labels on sample tubes intended for the 24position centrifuge rack
This information about the placement of barcode labels specifically refers to the usage of
24-position centrifuge racks and is an addition to the information in section 1.6.3 of the operator’s manual.
When a Hettich stand-alone centrifuge and 24-position centrifuge racks
(RD ID 06623611001) are used, make sure that:
•
No sample tubes that are covered with more than 3 layers of label are used.
•
Labels run the length of the tube and are parallel to the tube edge.
•
The labels are applied smoothly, without dog-eared corners, wrinkles, or edges that stick out.