- Home
- Home security & automation
- Carbon monoxide (CO) detectors
- Dräger
- Polytron 7000
- Operating instructions
advertisement
Dräger Polytron 7000
Approved as type P3U and type P3FB
Instructions for use
Dräger Polytron 7000
Contents
Contents
8VHLQDUHDVVXEMHFWWRH[SORVLRQKD]DUGV
'HWHFWLRQDQGPHDVXUHPHQWRIR[\JHQLQOLQHZLWK(1
'HWHFWLRQDQGPHDVXUHPHQWRIR[\JHQLQOLQHZLWK
&KHFNLQJWKHGLVSOD\IRUSURSHUIXQFWLRQLQJ
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Contents
&RQQHFWLQJWKHPHDVXUHPHQWPRGXOHZLWKWKHGRFNLQJ
$ODUPGHYLFHVRQO\ZLWKUHOD\PRGXOH
,QIRUPDWLRQDERXWWKHLQVWDOODWLRQRIWKHDVSLUDWLRQOLQH
3HUIRUPLQJDXWRPDWLFFDOLEUDWLRQ
6ZLWFKLQJDODUPDFNQRZOHGJHPHQWRQRURIIDWWKHJDV
7HVWLQJWKHDODUPUHOD\DQGWKHIDXOWUHOD\
5HVHWWLQJWKHPHDVXUHPHQWPRGXOHWRWKHIDFWRU\VHWWLQJV
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Contents
,QIRUPDWLRQDERXWWKHZDUQLQJVLJQDO
6ZLWFKLQJWKHZDUQLQJVLJQDORQRII
6HWWLQJWKHHOHFWULFFXUUHQWYDOXHRIWKHZDUQLQJVLJQDO
6HWWLQJWKHVLJQDOW\SHRIWKHPDLQWHQDQFHVLJQDO
$GMXVWLQJWKHHOHFWULFFXUUHQWRIWKHVWDWLFPDLQWHQDQFH
$GMXVWLQJWKH]HURVKLIWRIWKH P$VLJQDO
&RQILJXULQJWKHDGGUHVVIRUWKH3URILEXV 3$)RXQGDWLRQ )LHOGEXV
$FWLYDWLQJWKHVHQVRUUHSODFHPHQWIXQFWLRQ
(QDEOLQJGLVDEOLQJWKHDXWRPDWLFFDOLEUDWLRQIXQFWLRQ
6HQVRU WHVW RQO\ ZLWK VHQVRUWHVW GRQJOH RU ZLWK GLDJQRVWLF GRQJOH
3HUIRUPLQJWKHVHQVRUVHOIWHVWPDQXDOO\
5HVHWWLQJWKHVHQVRUWRWKHIDFWRU\VHWWLQJV
$GMXVWLQJWKHFDOLEUDWLRQLQWHUYDO
'DWD VWRUDJH VHWWLQJ RQO\ ZLWK WKH GDWD GRQJOH
,QIRUPDWLRQDERXWWKHPHDVXUHGYDOXHORJJHU
,QIRUPDWLRQDERXWWKHHYHQWORJJHU
2SHUDWLQJFRQGLWLRQVZLWK'U¦JHU6HQVRU2 DQG2 /6
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Contents
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Safety-related information
1.1
1.2
General safety notes
– Before using this product, carefully read the instructions for use.
– Strictly follow the instructions for use. The user must fully understand and strictly observe the instructions. Use the product only for the purposes specified in the
Intended use section of this document.
– Do not dispose of the instructions for use. Ensure that they are retained and appropriately used by the product user.
– Only trained and competent users are permitted to use this product.
– Comply with all local and national rules and regulations associated with this product.
– Only trained and competent personnel are permitted to inspect, repair and
service the product as detailed in these Instructions for Use (see chapter 8).
Further maintenance work that is not detailed in these Instructions for Use must only be carried out by Dräger or personnel qualified by Dräger.
Dräger recommends a Dräger service contract for all maintenance activities.
– Only use genuine Dräger spare parts and accessories when performing maintenance work, or the proper functioning of the product may be impaired.
– Do not use a faulty or incomplete product. Do not modify the product.
– If a pump module is used, care must be taken when opening the gas detector.
Through an existing glass tube there is risk of breakage and injury.
– If a pump module is used, do not block the lower gas inlet of the pump module.
This can destroy the sensor.
Use in areas subject to explosion hazards
– Equipment and components which are used in explosion-hazard areas and which have been inspected and approved in accordance with international or
European explosion-protection regulations may be used only under the specified conditions.
– In applications, which require devices of category 1G (zone 0) resp. EPL Ga, intensive electrostatic charging processes have to be prevented.
– Not for use in oxygen enriched atmospheres, i.e. more than 21 Vol% oxygen.
– If fitted at the time of delivery, or retrofitted at a later stage, with the relay module and/or the pump module, the device loses its explosion protection approval.
Operation of the device with the pump module and/or the relay module installed is not permitted in explosion-hazard areas.
– When measured gas is aspirated from explosion-hazard areas, suitable explosion protection measures are required, e.g. use of flame arresters.
– The relevant regulations for routing and connecting of power and signal lines to gas detection systems must be strictly adhered to.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
1.4
2
2.1
Conventions in this document
– The electric network (gas detector and alarm devices) in which the gas detector is installed must be equipped with a circuit breaker or a fuse.
– The circuit breaker or fuse must be easily accessible and must be marked as belonging to the gas detector.
– Observe the operational temperature range of the cables. Electric strength of the cable insulation corresponding to the supply for the alarm devices that are connected to the relays.
– If the gas detector is not connected to a controller, an alarm signalling device must be connected to the fault relay.
Detection and measurement of oxygen in line with EN
50104
– P3U is certified with the EU type examination certificate BVS 03 ATEX E 406 X and the type examination certificate PFG 16 G 005 X for the measurement of oxygen.
– The certified measuring function concerns the measuring range 0 to 25 % O 2
(v/v) for the purpose of measuring inertisation (BVS 03 ATEX E 406 X) plus the monitoring of the ambient air for oxygen deficiency and surplus
(PFG 16 G 005 X).
– Only configure alarms as “non-acknowledgeable”.
– Only use the readouts on the 4-20 mA analogue output or the switching states of the relays for purposes relating to safety.
– Approved sensors and components:
– DrägerSensor O 2 LS (6809630)
– DrägerSensor O 2 (6809720)
– Remote sensor (8317275)
– Relay module (8317360)
Conventions in this document
Meaning of the warning notes
The following alert icons are used in this document to provide and highlight areas of the associated text that require a greater awareness by the user. A definition of the meaning of each icon is as follows:
Alert icon Signal word
WARNING
CAUTION
Consequences in case of nonobservance
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in injury.
It may also be used to alert against unsafe practices.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Description
Alert icon Signal word
NOTICE
Consequences in case of nonobservance
Indicates a potentially hazardous situation which, if not avoided, could result in damage to the product or environment.
– HART ® is a registered trademark of the HART Communication Foundation.
– PROFIBUS
®
is a registered trademark of the PROFIBUS Nutzerorganisation e. V.
– FOUNDATION TM is a registered trademark of the Fieldbus Foundation.
3 Description
Polytron
5
4
3
2
M
OK
1 Control panel with display
2 Sensor
3 Bayonet ring
4 Docking Station
5 Measurement module
1
The Dräger Polytron 7000 is a stationary gas detector and is used for the monitoring of toxic gases and oxygen. The device converts the measured gas concentration into an electrical signal for further processing.
Frequency of the calculation of the measured value: 1 x per second (updating of the display, the 4-20-mA interface and the relays).
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Description
Ex works, the gas detector is configured with different interfaces and accessories.
3.3.1 Sensors
The gas detector is intended for the use of the DrägerSensor EC (electrochemical).
The measuring function in line with EN 50104 is certified with the oxygen sensors
O 2 (6809720) and O 2 LS (6809630).
3.3.2 Interfaces
Interface Type
4-20 mA
4-20 mA/HART
PROFIBUS PA
FOUNDATION Fieldbus H1
LON (Local Operating
Network)
P3U
P3U
P3FB
P3FB
–
Explosion protection approval
Yes
Yes
Yes
Yes
No
3.3.3 Accessories
Accessories Function
Pump module
(8317350)
Relay module
(8317360)
Daisy chain installation kit (8317282)
Remote sensor
(8317275)
Duct and pipe adapter
(Duct mounting kit)
Duct adapter for remote sensor
(8317617)
Calibration adapter V
(6810536)
Calibration adapter AC
(6809380)
Pump adapter
(8317976) for AC sensor
Aspiration of measured gas
Local switching of actuators
Connection of multiple gas detectors to one bus line (multidrop installation)
Installation of the sensor at a distance of up to 30 m from the measurement module.
Installation of the gas detector to a pipe in order to measure the gas concentration inside it.
Installation of a remote sensor to a pipe in order to measure the gas concentration inside it.
Supply of test gas
Supply of test gas together with the pump adapter (8317976)
Gassing of an AC sensor by means of the pump integrated in the gas detector
Explosion protection approval
No
No
Yes
Yes
Yes
Yes
No
No
No
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Description
Accessories
Dongles
Sensor-test dongle (8317619)
Diagnostic dongle
(8317860)
Data dongle
(8317618)
Function Explosion protection approval
Activation of additional functions
Sensor test functions that ensure the reliability and functionality of the sensor and of the gas warning device
Sensor diagnostics feature
(including sensor tests) to determine the strain and the remaining sensor life time
Data and event logging feature – stores measured values and events such as alarms and faults.
Enables graphic plotting of the measured values on a 15-minute time axis.
Yes
Yes
Yes
For information about additional accessories refer to the corresponding Dräger product information.
In combination with the integrated DrägerSensor, the gas detector is intended for continuous monitoring of gas concentrations.
For installation in:
Explosion-hazard area
Category
UL-/CSA area
Zones 0, 1, 2; Mining
1G, 2G, 3G; M1
Class I & II; Div. 1 & 2
3.5 Approvals
Depending on the interface, the gas detector is approved under the type
designation P3U or P3FB (see section 3.3.2 Interfaces).
3.5.1 ATEX
Device marking in accordance with 2014/34/EU
Types P3U & P3FB
BVS 03 ATEX E 406 X
0158 II 1G / I M1
Ex ia IIC T4 Ga -40 °C ≤ Ta ≤ +65 °C
(-40 °F ≤ Ta ≤ +149 °F)
Ex ia IIC T6 Ga -40 °C ≤ Ta ≤ +40 °C
(-40 °F ≤ Ta ≤ +104 °F)
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Description
Ex ia I Ma -40 °C ≤ Ta ≤ +65 °C
(-40 °F ≤ Ta ≤ +149 °F)
0158 II 3G
Ex ic IIC T4 Gc -40 °C ≤ Ta ≤ +65 °C
(-40 °F ≤ Ta ≤ +149 °F)
Ex ic IIC T6 Gc -40 °C ≤ Ta ≤ +40 °C
(-40 °F ≤ Ta ≤ +104 °F)
Power supply: U i = 30 V, I i = 0.3 A, P i = 700 mW, C i = 5 nF, L i = 50 μ H
Year of manufacture (via serial number) 1)
Dräger Safety, 23560 Lübeck, Germany
Type P3U
Safety-related characteristic values for the supply and signal circuit (outer terminals of the docking station):
Ui = 30 V, I i = 0.3 A, P i = 700 mW, C i = 5 nF, L i = 50 μ H
Type P3FB
FISCO bay unit
Power supply: U i = 24 V, I i = 0.38 A, P i = 5.32 W, C i = 5 nF, L i = 10 μ H
3.5.2 IECEx
Types P3U & P3FB
Ex ia IIC T4 Ga
Ex ia IIC T6 Ga
Ex ia I Ma
Ex ic IIC T4 Gc
Ex ic IIC T6 Gc
-40 °C ≤ Ta ≤ +65 °C
(-40 °F ≤ Ta ≤ +149 °F)
-40 °C ≤ Ta ≤ +40 °C
(-40 °F ≤ Ta ≤ +104 °F)
-40 °C ≤ Ta ≤ +65 °C
(-40 °F ≤ Ta ≤ +149 °F)
-40 °C ≤ Ta ≤ +65 °C
(-40 °F ≤ Ta ≤ +149 °F)
-40 °C ≤ Ta ≤ +40 °C
(-40 °F ≤ Ta ≤ +104 °F)
IECEx BVS 04.0003X
Year of manufacture (via serial number)*
Dräger Safety, 23560 Lübeck, Germany
Type P3U
Power supply: U i = 30 V, I i = 0.3 A, P i = 700 mW, C i = 5 nF, L i = 50 μ H
1) Structure of the serial numbers: The third letter of the serial number indicates the year of manufacture: M = 2019, N = 2020, P = 2021, R = 2022, S = 2023, T = 2024, U = 2025,
W = 2026, X = 2027, Y = 2028, Z = 2029 etc. (letters G, I, O, Q are omitted)
Example: Serial number ARMB-0001: The third letter is M, meaning that the device was manufactured in 2019.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Description
Type P3FB
FISCO bay unit
Power supply: U i = 24 V, I i = 0.38 A, P i = 5.32 W, C i = 5 nF, L i = 10 μ H
3.5.3 UL
Types P3U & P3FB
Only in accordance with intrinsic safety classified for operation in explosion-hazard areas.
Class I, Div. 1&2, Groups A, B, C, D
Class II, Div. 1&2, Groups E, F, G
Use in accordance with Dräger control drawing SE20105.
T4
T6
-40 °C ≤ Ta ≤ +65 °C
(-40 °F ≤ Ta ≤ +149 °F)
-40 °C ≤ Ta ≤ +40 °C
(-40 °F ≤ Ta ≤ +104 °F)
Type P3U
Power supply: V max = 30 V, I max = 0.3 A, P i = 700 mW, C i = 5 nF, L i = 50 μ H
Type P3FB
FISCO bay unit
Power Supply: V max = 24 V, I max = 0.38 A, P max = 5.32 W, C i = 5 nF, L i = 10 μ H
3.5.4 CSA
NRTL/C
® LR97594
Types P3U & P3FB
Intrinsic safety
Class I, Div. 1, Groups A, B, C, D
Class II, Div. 1, Groups E, F, G
Use in accordance with Dräger control drawing SE20106.
Type P3U
Power supply: V max = 30 V, I max = 0.3 A, P max = 700 mW, C i = 5 nF, L i = 50 μ H
Type P3FB
FISCO bay unit
Power supply: V max = 24 V, I max = 0.38 A, P max = 5.32 W, C i = 5 nF, L i = 10 μ H
3.5.5 Detection and measurement of oxygen in line with EN 50104
The measuring function in line with EN 50104 is certified from software version 8.6 onwards.
Measuring function for explosion protection
BVS 03 ATEX E 406 X
Device configuration
Type P3U with DrägerSensor O 2 LS (6809630) or
O 2 (6809720)
Measuring range
0 to 25 vol.-% O 2
Purpose
Inertisation measurement
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Use
Measurement of oxygen
PFG 16 G 005 X
Device configuration
Type P3U with DrägerSensor O 2 LS (6809630) or
O 2 (6809720)
Measuring range
0 to 25 vol.-% O 2
Purpose
Detection of oxygen deficiency and oxygen surplus
The pump module is not part of the EU type examination certificate BVS 03 ATEX E
406 X and the type examination certificate PFG 16 G 005 X.
4 Use
In addition to the control panel, the unit can also be operated using, for example, the Dräger PolySoft software, FDT applications or manual control units.
If individual parameters were changed with a configuration software, all parameters from the device must be read back and verified manually.
4.1 Display
Example display
O2
20.9
Vol.%
O2
19.0
A1
Description
Normal operation – The display shows the gas concentration, the measured gas and the unit of measure.
Overshooting (in the case of O 2 also undershooting) of the alarm thresholds – The display in the example shows the pre-alarm A1.
Overshooting the full scale value specified for the sensor (see instructions for use/data sheet for the sensor)
4-20-mA interface: Full-scale value
Relay: A2 relay energised
O2 A2
No monitoring of the gas concentration and/or alarm state occur in the special state.
Special states on the gas detector exist in the following cases:
– Negative gas concentration
– Fault
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Use
– Disabled alarms (display readout and analogue/digital measured value readout remains active)
– Calibration procedures
– Warm-up phase
– Maintenance
Example display
O2
20.9
Vol.%
Description
Maintenance – When certain functions are invoked
(e.g. calibration, sensor replacement function), the display shows the symbol .
4-20-mA interface: Maintenance signal
Relay: Energised or not energised depending on function
General error – The display shows the symbol .
4-20-mA interface: Fault signal
Relay: Fault relay energised
O2
20.9
Vol.%
A warning is present. The display shows the symbol
4-20-mA interface: Measured value
1)
Relay: Fault relay energised (with a disabled alarm)
1) The warning signal is also issued for the measured value when a warning signal is selected.
Factory setting: Warning signal disabled.
Conduct during warm-up phase 1
4-20-mA interface: Maintenance signal
Relay: Fault relay energised
During warm-up phase 1 the display shows the time remaining until warm-up phase
1 is completed.
Conduct during warm-up phase 2
4-20-mA interface: Measured value
Relay: Fault relay not energised
During warm-up phase 2 the display shows the measured value.
Undershooting the measuring range (O 2 sensors only)
If the measured value drops below the measuring range set, the 4-20-mA interface emits 3.8 mA ... to 4 mA and the display shows the symbol .
If the measured value drops further, the gas detector behaves in the same way as for a general error. The 4-20-mA interface issues the fault signal, the display shows the symbol and the fault relay is energised.
Undershooting the measuring range is a non-latching error. If the measured value is again in the measuring range, the error is no longer issued.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
4.1.3
4.1.4
Use
Meaning of the symbols
Symbols on the right side of the displays indicate the device state. Some symbols are only displayed if the gas detector is operated with the corresponding component.
Symbol Explanation
A fault is present. Fault signal is issued.
A warning is present. A warning signal is issued.
1)
A maintenance signal is issued.
Pump is installed.
Pump flow error
The configured full scale value is overshot in the 4-20 mA interface
The configured threshold limit value is undershot in the 4-20mA interface.
Polling address of the analogue interface is set to a fixed value
(Multidrop operation) and transmits via HART.
Sensor diagnostic function – The sensor is ready for operation.
Sensor diagnostic function – The sensor is ready for operation but nearing the end of its operating lifetime.
Sensor diagnostic function – The sensor is ready for operation but should be replaced soon.
The buffer mode of the data logger is set to overwriting (roll).
The buffer mode of the data logger is set to holding (stack).
1) If the device is in maintenance mode (e.g. warm-up phase 1), the maintenance signal is issued instead of the warning signal.
LED indication for LON version
The LON version of the gas detector has three LEDs located behind the display.
LED display Meaning
The green LED is illuminated
The gas detector is functioning properly.
The green LED is flashing
A warning is present.
Orange LED The gas detector communicates with the control unit via LON.
Red LED An error is present.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Use
4.1.5 Checking the display for proper functioning
In order to recognise defective pixels in the display, it is possible to invoke the
Show Fault Codes menu image.
The menu image is suited since it shows many contiguous pixels, making it possible to recognise line, column or individual pixel faults. Displayed fault codes are not relevant in this context.
1. Open the menu (see chapter 4.4.3).
2. In the menu, select Information > Instrument > Show Fault Codes .
3. Check if all pixels are displayed.
4.3
Polytron
1
2
M
OK
1 Up key or function key
2 Down key or menu key
3 OK key
3
Info mode and function key
In measuring mode, the Info mode can additionally be activated. It provides an overview of device information presented on multiple pages.
Example display
Page 1
07.11.2011 12:34
SW Version : 8.0
Part No. : 8317778
Serial No.
: ARUA0001
DeviceCode: 00006317
1/4
Instrument Info
Page 2
Sensorname: O 2
Part No. : 6809630
Serial No.
: XXXXXXX
EEPROM Typ : 1
EEPROM Vers.: 1
2/4
Sensor Info
Description
Device information
Line 1 – Date and time
Line 2 – Software version
Line 3 – Device part no.
Line 4 – Device serial no.
Line 5 – Device code
Sensor information
Line 1 – Sensor name
Line 2 – Sensor part no.
Line 3 – Sensor serial no.
Line 4 – EEPROM type
Line 5 – EEPROM version
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Use
4.3.2
4.3.3
4.3.4
Example display
Page 3
Gasname : O 2
Range : 25.00 Vo
4–20 SP : 25.00 Vo
Alarm A1 : 19.00 Vo
Alarm A2 : 23.00 Vo
3/4
Sensor Config.
Page 4
2)
Description
Sensor settings
Line 1 – Gas name
Line 2– Measuring range (cannot be edited) and unit of measurement
Line 3 – Measuring range for the analogue interface
Displayed only if the analogue interface card is installed
Line 4 – Alarm threshold A1 and unit
1)
Line 5 – Alarm threshold A2 and unit
1)
Pump information
Line 1 – Pump output
Line 2 – Flow threshold for error
Line 3 – Flow threshold for warning
Line 4 – Running time of the pump
Pump info
4/4
1) Displayed only if a relay module is used!
2) Displayed only if a pump module is used!
Activating Info mode
● In measuring mode, press and hold down for approx. 3 seconds.
Navigation in Info mode
Key
M
OK
Function
Scrolls to the next page.
Scrolls to the previous page.
Exits Info mode.
After 30 seconds without key operation, the gas detector automatically exits the Info mode.
Using the function key
The function key can be used to directly invoke a function that has been assigned to the function key. By default, the error log is displayed.
● In measuring mode, press and hold down for approx. 1 second.
For configuring of the function key (see section 9.7 Configuring the function key).
4.4 Menu
Key
M
Function
Navigates upward.
Sets values.
Navigates downward.
Sets values.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Use
Key
OK
Function
Confirms inputs.
Selects menus and functions.
4.4.2 Passwords
Calibration password Access to
– Information menu
Settings password
– Calibration menu
Factory setting: _ _ _ 1
Access to
– Information menu
– Calibration menu
– Settings menu
Factory setting: _ _ _ 2
For setting the passwords, see section 9.3.
4.4.3
4.4.4
Opening the menu
● To open the Information directly: In measuring mode, press and hold down
M for approx. 1 second.
● To open the Information and Calibration menus: a. In measuring mode, press and hold down for approx. 3 seconds.
M b. Select Enter password .
c. Enter the calibration password.
● To open all menus: a. In measuring mode, press and hold down for approx. 3 seconds.
M b. Select Enter password .
c. Enter the settings password.
Reading out information
In the menu Information choose the appropriate menu item:
Instrument
Show Notice
Show Fault
Show Fault
Codes
Shows existing warnings. For troubleshooting with respect to the warning number displayed
If multiple warnings are present, this is indicated (e.g. 1/3 = page 1 of 3.).
Shows existing errors. For troubleshooting with respect to
the error number displayed (e.g. #100) see section 7.1.
If multiple errors are present, this is indicated (e.g. 1/3 = page 1 of 3.).
Shows present error and warning codes in a table. If all numerical groups are displayed with "00", no error is present.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
5
5.1
5.1.1
Installation & commissioning
Show Module
Sensor
Vitality
1)
Show sensortemp.
Datalogger
Shows the remaining sensor vitality in %.
Dräger recommends to replace the sensor when the vitality
< 25 %.
Last Cal. Date Shows the date of the last calibration.
Next Cal. Date Shows the due date for the next calibration.
Shows the current and the max. measured sensor temperature values.
Datalogr. status
2)
Show graph
2)
Shows the installed hardware modules. To show detailed information, select module in the menu.
= The module is installed
= The module is not installed
The detailed information also shows the software version of the module which is used.
Shows for the data logger and for the event logger respectively whether they are switched on or off.
Represents the measured values on a time axis of 15 minutes.
1) Only works with the diagnostic dongle
2) Only works with the data dongle
Installation & commissioning
Mounting the measurement module
Preparing the docking station
Prerequisite:
– Docking station is mounted and wired.
(as per assembly instructions Dräger Docking Station for Polytron
®
3000/7000)
1. Remove the dust and rain protector, if in place.
2. Inspect the sealing of the docking station for contamination, clean if necessary.
3. From the inside, check the position of the eccentric screws.
The screws must be in engaged position, with the eccentric opening pointing upward.
4. Using a 5 mm Allen key (without spherical head), correct the position of the eccentric screws, if necessary.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Installation & commissioning
5.1.2 Connecting the measurement module with the docking station
NOTICE
Ingress of humidity
If the measurement module is not flush with the docking station, humidity can enter, damaging the gas detector.
► Make sure that the measurement module is correctly seated in the docking station.
1. Horizontally slide the measurement module into the docking station and lower it.
The front of the measurement module is flush with the docking station. A gap of approx. 3 mm (0.12") remains at the top between docking station and measurement module.
5.2
2. Using the 5 mm Allen key (without spherical head), turn the eccentric screws approx. 180° in clockwise direction towards .
The gap at the top between docking station and measurement module closes.
The measurement module is locked.
Installing the sensor
If the gas detector is already switched on and an existing sensor is to be
Prerequisite:
– Measurement module is installed.
1. Loosen the hexagon socket screw at the bayonet ring.
2. Unscrew the bayonet ring from the gas detector.
3. Remove the blank disk.
4. With the Dräger logo pointing forward, insert the sensor into the sensor opening.
5. Screw the bayonet ring back on.
6. Tighten the hexagon socket screw at the bayonet ring.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
5.3
5.3.1
5.3.2
Installation & commissioning
Alarm devices (only with relay module)
If alarm devices are to be operated, the electrical connection of the docking station must be in 3-wire technology (see assembly instructions Dräger Docking Station for
Polytron ® 3000/7000).
Information about connecting
Three potential-free relay outputs are available at the relay module:
– alarm relay A1 (pre-alarm)
– alarm relay A2 (main alarm)
– fault relay (fault signal)
Pin-out of the relay plug
1
2
4
3
1 normally closed
2 common
3 normally open
4 not connected
Connecting alarm devices
WARNING
Risk of electric shock
Potential differences can lead to insulation faults.
► Do not combine electrical loads with different voltage types (AC or DC).
► If DC loads are used, make sure that only devices with identical rated DC voltages are controlled via the relay contacts (e.g. ≤ 120 V).
► If AC loads are used, make sure that only the relay contacts are only connected to devices with which they share one phase.
WARNING
Risk of electric shock
Improper use can lead to personal injury.
► Before plugging in or unplugging the relay plug, disconnect alarm devices from the power supply.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Installation & commissioning
Equipment
– connection cable diameter of the connection cable: 6 to 9.5 mm diameter of the individual wires: 3 mm max.
cable core cross-section: 2.5 mm
2
(AWG 14)
– ferrite sleeve
– ferrules
– cable bushing (order no. 18 90 086)
5.4
5.4.1
Procedure:
1. Mount the cable bushing with the ferrite sleeve to the connection cable. Screw all connections tight.
2. Make a note the pin-out of the relay outputs on the label of the relay cap.
3. Insert and lock the connection cable.
4. Close the relay cap.
Pump adapter (only with pump module)
If the pump module is to be operated, the electrical connection of the docking station must be in 3-wire technology (see assembly instructions Dräger Docking
Station for Polytron
®
3000/7000).
Mounting the pump adapter
Polytron
OK
2
1
1. Loosen the hexagon socket screw at the bayonet ring.
2. Unscrew the bayonet ring (2) from the gas detector.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Installation & commissioning
5.4.2
5.5
3. If necessary, insert the sensor.
4. Place the mounting ring (1) of the pump adapter over the sensor opening.
5. Screw the bayonet ring back on.
6. Tighten the hexagon socket screw at the bayonet ring.
7. Observe the installation direction of the pump adapter.
The installation direction certain the direction of gas flow between pump and sensor (mode of operation).
Mode of operation Description
Grommets for supply and waste air point to the left, symbol points to the front
The pump is upstream of the sensor in the gas flow; the sensor is on the overpressure side of the pump.
This is the preferred mode of operation for all sensors.
Grommets for supply and waste air point to the right, symbol points to the front
The pump is downstream of the sensor in the gas flow; the sensor is on the negative pressure side of the pump.
This mode of operation should be chosen only in justified cases.
For the DrägerSensor O 2 LS (6809630) the mode of operation is only permissible up to a maximum flow rate of 0.5
L/min.
This mode of operation is not permitted for the Dräger-
Sensor O 2 (6809720).
8. Insert the bushings of the pump adapter into the bore holes in the underside of the docking station.
The sealing slides over the sensor.
9. Turn the mounting ring in clockwise direction until the pump adapter is secured.
Information about the installation of the aspiration line
The selection of the material of the aspiration hose/aspiration pipe and the aspiration distance influence the response time of the measured signal. In the worst case, reaction with the material used or absorption can result in no measurable gas concentration reaching the sensor.
A pressure difference between the aspiration location and the environment of the gas detector can lead to an additional measuring error. The maximum permissible pressure difference is 50 mbar.
Contact Dräger for information about selecting a suitable hose/pipe material.
Commissioning the gas detector
Prerequisite:
– The sensor is installed.
1. Switch on the power supply.
The gas detector starts the warm-up phase.
The time remaining until the sensor reaches operational readiness. The maintenance signal is issued. If a relay module is installed, the fault relay is energised.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Calibration
2. Wait until the warm-up phase is completed.
Depending on the installed sensor, the warm-up phase can take between
5 minutes and 12 hours.
It is already possible to make settings in the menu:
3. Calibrate the sensor (see section 6).
4. Perform tests:
● Check signal transmission to the control unit.
● Test the alarm relay and the fault relay (only with relay mode, see
5. If a pump module is used:
● Check the aspiration path for tightness.
● Compare the flow rate at the aspiration location and downstream of the gas detector.
6 Calibration
Calibration involves adjusting the zero-point and the sensitivity of the sensor.
6.2
Depending on the type of calibration, different test gases are used.
Zero gas
Zero gas is a test gas used for zero-point adjustment. The zero gas used can be fresh air from the ambient atmosphere. The fresh air must be free from the measured gas or other substances that have cross-sensitivity. Nitrogen (N 2 ) is used in the case of O 2 sensors.
Target gas
Target gas is a test gas used for sensitivity adjustment. The target gas used can be the measured gas or a surrogate gas. In the case of O 2 sensors, no target gas is needed, and the oxygen from the ambient air (fresh air) is used.
Preparing the test gas supply
The mode of test gas supply depends on whether the gas detector is equipped with a pump module or not.
Prerequisite:
– The warm-up phase of the sensor is completed (see section 5.5 Commissioning the gas detector).
– Date and time have been set (see section 9.4).
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Calibration
WARNING
Health hazard from test gas
Breathing in of test gas can be harmful to health or lead to death.
► Do not inhale the test gas.
► Observe risks connected with the test gas, hazards notes and safety advice (see for example safety data sheets, instructions on the testing media).
CAUTION
Alarms triggered by test gas
Test gas that is still present can trigger alarms.
► Make sure that the supply of test gas is ended.
6.2.1
2SWLRQ
Gas detector without pump module
The test gas must be fed without being pressurised. This corresponds, e.g. if a test gas cylinder is used, a flow rate of 500 mL/min. If test gas ampoules are used, unpressurised supply is automatically the case.
Using a test gas cylinder
Equipment:
– test gas cylinder with pressure reducer, in the case of aggressive gases with stainless steel pressure reducer
– calibration adapter with hose nozzles (order no. 68 10 536)
– hose, type appropriate to the gas properties (e.g. FKM hose order no. 12 07 068)
1. Connect hose to calibration adapter and test gas cylinder.
2. To discharge the test gas, connect a hose to the second connection of the calibration adapter.
3. Plug the calibration adapter onto the sensor.
Polytron
OK
2SWLRQ Using test gas ampoules
Test gas ampoules are only used for the sensitivity adjustment for the target gas.
Equipment:
– calibration flask (order no. 68 03 407) with the corresponding adapter
– Test gas ampoule
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Calibration
● Plug the calibration flask onto the sensor.
OK
68 07 407
6.2.2
2SWLRQ
Gas detector with pump module
If a pump module is used, equipment is required that makes it possible to feed the unpressurised test gas into the volume flow of the pump.
Using a gas bag
Equipment:
– Gas bag
– Test gas cylinder
– hose, type appropriate to the gas properties (e.g. FKM hose order no. 12 07 068)
● Connect hoses to the pump module.
Do not connect the hose to the gas bag until calibrating.
OK
+
2SWLRQ Using test gas cylinder and flowmeter
Equipment:
– test gas cylinder with pressure reducer, in the case of aggressive gases with stainless steel pressure reducer
– hose distributor
– flowmeter
– hose, type appropriate to the gas properties (e.g. FKM hose order no. 12 07 068)
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
● Connect hoses. Integrate flowmeter with hose distributor.
OK
+
Calibration
2SWLRQ Using a test gas ampoule
Test gas ampoules are only used for the sensitivity adjustment for the target gas.
Equipment:
– calibration flask (order no. 68 03 407)
– adapter (order no. 68 04 620)
– Test gas ampoule
– hose, type appropriate to the gas properties (e.g. FKM hose order no. 12 07 068)
1. Plug the adapter onto the calibration flask.
2. Connect hoses.
OK
68 07 407
68 04 620
+
6.3 Adjusting the zero-point
Prerequisite:
– The warm-up phase of the sensor is completed (see section 5.5).
– Test gas supply is prepared (see section 6.2).
Zero adjustment can be aborted at any time. To abort during calibration, select
Previous .
If no entries are made, the gas detector automatically switches over to measuring mode after 15 minutes.
Procedure, using an EC-H 2 S sensor as an example
1. Select Calibration > Zero calibration > EC-H 2 S .
The 4-20 mA interface issues the maintenance signal.
2. Supply zero gas or fresh air, respectively:
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Calibration
Option Description
Test gas cylinder Set the pressure reducer to a flow rate of 500 mL/min.
Gas bag
Test gas cylinder and flowmeter
Fill the gas bag using the test gas cylinder. Connect hose leading to the pump module with gas bag.
Slowly open the pressure reducer of the test gas cylinder until the flowmeter indicates a flow.
3. Select Next .
The current value and the gassing time are displayed.
In the case of EC-O 2 sensors, only checking of the zero-point value is performed instead of zero adjustment. The zero-point value must be less than
0.6 volume percent. If this is not the case, an error is set in the sensor.
4. When the displayed measured value has stabilised (max. wait time = 3 minutes), select calibrate .
5. End the zero gas supply.
6. Select Back to menu .
Prerequisite:
– Zero adjustment has been performed (see section 6.3).
– Test gas supply is prepared (see section 6.2).
Span calibration can be aborted at any time. To abort during calibration, select
Previous .
If no entries are made, the gas detector automatically switches over to measuring mode after 15 minutes.
Procedure, using an EC-H 2 S sensor as an example
1. Select Calibration > Span calibration > EC-H 2 S .
The 4-20 mA interface issues the maintenance signal.
The values for the span calibration using the target are displayed.
Example:
" Cal.gas : H 2 S
Cal.unit : ppm
Concentr.: 000025 "
If the displayed values do not match the target used, use the control keys to adjust the values to the target gas.
2. Select Next
3. Supply target gas:
Option Description
Test gas cylinder Set the pressure reducer to a flow rate of 500 mL/min.
Test gas ampoule Break the test-gas ampoule in the calibration flask.
Gas bag Fill the gas bag using the test gas cylinder. Connect hose leading to the pump module with gas bag.
Test gas cylinder and flowmeter
Slowly open the pressure reducer of the test gas cylinder until the flowmeter indicates a flow.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
6.5
Calibration
4. Select Next .
The current value and the gassing time are displayed.
5. When the displayed measured value has stabilised (max. wait time = 3 minutes), perform the calibration.
a. Select calibrate .
The target value, the current value and the sensor vitality (bar diagram) are displayed.
b. Select Accept value .
c. End the target gas supply.
The calibration interval and the date for the next calibration are displayed.
d. Select Back to menu .
Performing automatic calibration
Automatic calibration involves zero adjustment and span calibration in a combined process. Automatic calibration can be used alternatively instead of separate zero adjustment and span calibration.
Requirements:
– The automatic calibration function is activated (see section 11.2).
– The warm-up phase of the sensor is completed (see section 5.5).
– Test gas supply is prepared (see section 6.2).
Procedure, using an EC-O 2 sensor as an example
The process of automatic calibration for EC-O 2 sensors is shorter.
1. Select Calibration > Auto calibration .
The 4-20 mA interface issues the maintenance signal.
" Fresh air cal.
is running!
" is displayed.
Instead of performing zero adjustment, the zero-point value is checked. After that, the sensitivity is calibrated.
If the automatic calibration was successful, the message " EC-O 2 : ok " is displayed.
2. Select Back to menu .
Procedure, using an EC-H 2 S sensor as an example
1. Select Calibration > Auto calibration .
The 4-20 mA interface issues the maintenance signal.
The zero-point is adjusted. " Fresh air cal.
is running!
" is displayed.
If the zero adjustment was successful, span calibration is performed.
The values for the span calibration using the target are displayed.
Example:
" Cal.gas : H 2 S
Cal.unit : ppm
Concentr.: 000025 "
If the displayed values do not match the target used, use the control keys to adjust the values to the target gas.
2. Supply target gas.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Troubleshooting
Option Description
Test gas cylinder Set the pressure reducer to a flow rate of 500 mL/min.
Test gas ampoule Break the test-gas ampoule in the calibration flask.
Gas bag Fill the gas bag using the test gas cylinder. Connect hose leading to the pump module with gas bag.
Test gas cylinder and flowmeter
Slowly open the pressure reducer of the test gas cylinder until the flowmeter indicates a flow.
3. Select Next .
The current measured value and the gassing time are displayed.
4. Wait until the measured value has stabilised. When the displayed measured value has stabilised (max. wait time = 3 minutes), select calibrate .
With some sensors, the gas detector recognises by itself if the measured value has stabilised. If one of this sensors is installed, the gas detector starts the calibration automatically, without the need to select calibrate manually.
The target value and the current value are displayed.
5. End the target gas supply.
6. When the current value is equal to the target value, select Accept value .
7 Troubleshooting
If the display is without function, have the gas detector checked by Dräger.
The fault or warning numbers listed below are displayed in the menu (see
7.1 Fault
Number
# 1
# 2
# 61
# 63
# 64
# 65
# 67
Cause
Serious data error in the measurement module – various causes.
Serious device error – various causes.
Data error on the interface card – various causes
Hardware or software fault in the pump module.
Flow rate of the pump module undershoots the error threshold.
Reliable measurements are no longer possible.
Remedy
Reset the measurement module to the
factory settings (see section 9.8).
If the error persists, have the gas detector checked by DrägerService.
Have the gas detector checked by
DrägerService.
Have the gas detector checked by
DrägerService.
Replace the pump module.
Check hoses for blockages, re-adjust the flow rate, if necessary.
3-wire cable interrupted.
Check connections
Relay module not contacted correctly.
Check the connector of the relay module or re-plug.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Troubleshooting
Number
# 100
# 101
# 102
# 103
# 106
# 107
# 108
# 109
# 121
# 125
# 129
# 130
# 134
# 136
Cause
Measurement module cannot detect a sensor.
Sensor data error in the gas detector.
Gas detector does not support the sensor version.
Sensor data error in the gas detector.
Zero adjustment failed
Span calibration failed
Sensor data error
Device error
Remedy
Remove and re-install the sensor (see
If the error persists, check the sensor connector or replace the sensor.
Remove the sensor and install it again.
If the error persists, have the gas detector checked by DrägerService.
Use a compatible sensor as per spare parts list.
Reset the sensor to the factory set-
If the error persists, have the gas detector checked by DrägerService.
Perform zero adjustment again (see
Perform span calibration again (see
Replace the sensor.
Check sensor contacts, otherwise have the gas detector checked by
DrägerService.
Perform automatic calibration again
(see section 6.5), making sure that the
ambient air is free from foreign gas.
Replace the sensor.
Fresh air calibration failed
(1st step of automatic calibration)
Sensor not ready for operation.
Sensor electrolyte has evaporated.
Sensor lock is switched on.
A sensor with a different part number than previously configured is installed.
Contact between sensor and measurement module is faulty.
Sensor hardware fault.
Replenish sensor electrolyte.
Switch off the sensor lock (see
section 11.4) or use a new sensor with
the same part number.
Check sensor contacts. Remove and re-install the sensor several times.
If the error persists, replace the sensor.
Remove and re-install the sensor. If the error persists, replace the sensor.
7.2 Warnings
Number
# 1
Cause
Data error in the gas detector – various functions (e.g. for data storage) can be impaired.
Remedy
Reset the gas detector to the factory
If the error persists, have the gas detector checked by DrägerService.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Troubleshooting
# 106
# 111
# 112
# 114
# 115
Number
# 51
# 52
# 53
# 58
# 59
# 64
# 119
# 120
# 131
# 132
# 135
Cause
Data logger in hold mode
(Stack) is filled to 100 %. It is no longer recording data.
Data logger in hold mode
(Stack) is filled to 90 %.
Remedy
Read out the data, clear and restart
the data logger (see section 12).
Read out the data as soon as possible, then clear and restart the data logger.
No valid date or time is set. Set time and date (see section 9.4).
Dongle was removed without having been deactivated.
Deactivate Dongle prior to removing
Hardware fault of a dongle. Replace the dongle.
The pump is worn Replace the pump module.
Flow rate of the pump module undershoots the warning threshold. Reliable measurements will soon no longer be possible.
Zero deviation too high.
Check hose for blockage, re-adjust the flow rate, if necessary.
Sensor not operated in the specified temperature range.
Perform zero adjustment (see
Operate sensor in the specified temperature range.
Sensor near end of life.
Calibration interval expired.
The sensor was exposed to an excessively high gas concentration for a long period.
The sensor is not yet fully warmed up. Measuring errors can occur.
Replace the sensor (see section 8.2).
Calibrate the sensor again (see
Reduce excessive gassing.
Wait until the sensor has fully warmed up.
The sensor was exposed to an excessively high gas concentration for a long period.
The sensor will soon no longer be ready for operation.
Sensor electrolyte has evaporated. Measurements will soon no longer be possible.
Information, e.g. the part number or the serial number, is not available.
Reduce excessive gassing. If the problem persists, replace the sensor
Replace the sensor (see section 8.2).
Replenish sensor electrolyte.
Disconnect the gas detector from the mains and restart it.
If the error persists, have the gas detector checked by DrägerService.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Maintenance
8 Maintenance
8.1 Preparing a maintenance plan
The gas detector must be serviced at regular intervals. The intervals and activities are defined in the maintenance plan by the person who is responsible for the gas detection system.
The maintenance interval recommended by Dräger is 6 months.
Dräger recommends the following activities:
– Perform tests:
– Check signal transmission to the controller (see section 10.2.9).
– Test the alarm relay and the fault relay (only with relay module, see
– Checking the display for proper functioning 4.1.5
– Service the sensor (e.g. replace the selective filter).
– Calibrate the sensor (see section 6 Calibration).
– If a pump module is used:
– Check the aspiration path for tightness.
– Compare the flow rate at the aspiration location and downstream of the gas detector.
– Carry out inspection at regular intervals in accordance with applicable national regulations (e.g. EN 60079-29-2, EN 45544-4, T021/T023).
8.2 Replacing the sensor
Even in explosion-hazard areas, the sensor can be replaced without interruption of the supply voltage.
1. If a sensor of a different type is installed, it may be necessary to switch off the
sensor lock (see section 11.4).
2. Activate the sensor replacement function (see section 11.1). Otherwise a fault
signal will be issued when the sensor is pulled off.
3. Loosen the hexagon socket screw at the bayonet ring.
4. Unscrew the bayonet ring from the gas detector.
5. Remove the old sensor.
6. With the Dräger logo pointing forward, insert the sensor into the sensor opening.
7. Screw the bayonet ring back on.
8. Tighten the hexagon socket screw at the bayonet ring.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Device settings
9.1
9.1.1
Pump (only with pump module)
Adjusting the pump output
Equipment:
– Pump adapter (order no. 8320900)
– flowmeter
1. Select Settings > Instrument > Pump > Pump flow .
The 4-20 mA interface issues the maintenance signal.
" Flowalarm woud be disabled . Please usea Flowmeter ." is displayed.
If a relay module is installed, the fault relay is energised.
2. Connect pump adapter and flowmeter.
OK
L/min
+
9.1.2
9.2
9.2.1
3. Select Next .
4. Adjust the pump output setting and check at the flowmeter. For a short response time select a high pump throughput rate.
5. Select Next .
The flow thresholds are displayed.
Flow threshold for error: 0.3 L/min
Flow threshold for notice: 0.4 L/min
Displaying the pump runtime
– Select Settings > Instrument > Pump > Operating time .
Alarms (only with relay module)
Switching the alarm on or off
1. Select Settings > Instrument > Alarm > Alarm on/off .
enable Alarm relay is switched on.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Device settings
9.2.2
9.2.3
disable Alarm relay is switched off.
No alarm is reported via the alarm relays and the HART interface.
The fault relay is energised to display the status of the disabled alarm relays. The display shows a warning symbol .
2. Select Enable or Disable .
Configuring the relay
1. Select Settings > Instrument > Alarm > Relay settings .
2. Make the settings.
Normally energized
Alarm erregt
In normal state (no alarm), a current flows through the coil of the relay (fail-safe).
The coil of the relay is flowed through by current in alarm state.
Overview of alarm settings
Alarm condition met
Alarm configuration
Acknowledging
Latching Non Acknowledging
Non Latching
Acknowledging
Non Acknowledging
Acknowledge key operated
Acknowledge key not operated
Alarm disabled
Alarm enabled
Alarm disabled
Alarm enabled
Alarm enabled
Alarm enabled
Alarm enabled
Alarm enabled
Alarm condition ceases
Acknowledge key operated
Acknowledge key not operated
Alarm configuration
Acknowledging
Latching Non Acknowledging
Non Latching
Alarm disabled
Alarm disabled
Alarm enabled
Alarm enabled
Acknowledging
Non Acknowledging
Alarm is disabled
Alarm is automatically disabled if the alarm condition ceases
Alarm is automatically disabled if the alarm condition ceases
Alarm is automatically disabled if the alarm condition ceases
1. Select Settings > Instrument > Alarm > Cfg. Alarm A1 or Cfg. Alarm A2 .
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Device settings
9.2.5
2. Make the settings. Use Next to call up the steps in turn.
a. Set the value for the alarm threshold.
Min./max. Alarm thresholds are dependent on the sensor.
Alarm thresholds for O 2 sensors can be set anywhere between 0.1 vol% and
25 vol%.
b. Set the alarm direction.
Rising
Falling
An alarm is issued if the measured value is above the alarm threshold.
An alarm is issued if the measured value is below the alarm threshold.
c. Set the latch mode.
If the gas detector is used to detect oxygen, Alarm A1 and Alarm A2 must be configured as latching. If both alarm thresholds have the same direction
(activation when the oxygen concentration is increasing or decreasing), the first alarm may be configured as Non Latching .
Latching
Non Latching
Alarm relays do not change their state automatically when the alarm condition ceases. Alarm must be acknowledged manually.
Alarm relays change their state when the alarm condition ceases. Alarm does not need to be acknowledged manually. The gas detector is no longer in alarm state.
d. Set the alarm acknowledge mode. For safety-relevant switching actions, configure the alarm as Non Acknowledging .
Acknowledging Alarm relays can be acknowledged although an alarm is active.
Non Acknowledging
Alarm relays can only be acknowledged when the alarm is no longer active.
e. Set the hysteresis value.
This feature prevents the relay from being energised and deenergised in rapid sequence.
Example: The alarm threshold was set to 40 ppm and the hysteresis to
3 ppm. The alarm remains active until the measured value has fallen below
37 ppm.
f. At the overview of the settings made, select Confirm .
Switching alarm acknowledgement on or off at the gas detector
Enables/disables the possibility to acknowledge alarms at the gas detector by
When this function is switched off and the alarm is configured as latching, an alarm can only be acknowledged by a voltage interruption to the gas detector.
1. Select Settings > Instrument > Alarm > Acknowledge mode .
2. Select Enable or Disable .
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
9.6
9.7
9.2.6
9.4
9.5
Device settings
Testing the alarm relay and the fault relay
During the relay test, an error state is simulated and the corresponding relay is energised.
1. Select Settings > Instrument > Alarm and the desired signal condition.
Set Alarm A1
Set Alarm A2
Set Fault
Simulate pre-alarm (alarm 1).
Simulate main-alarm (alarm 2).
Simulate fault signal.
2. Select Enable or Disable .
1. Select Settings > Instrument > Passwords and the password to be changed.
2. Set the password.
Setting date and time
1. Select Settings > Instrument > Date and time .
2. Select the line for date or time.
3. Make the settings.
Setting the time format
1. Select Settings > Instrument > Time format .
2. Select the line for the time format of date or time.
3. Select the time format (European or American).
Setting the language
1. Select Settings > Instrument > Language .
2. Select the language.
Configuring the function key
1. Select Settings > Instrument > Function key .
2. Select the function to be called.
Show graph
Show Fault
Show Notice
1) Displays the measured values of the sensor graphically on a time axis.
Shows existing errors.
Shows existing warnings.
Show Fault Codes Shows fault codes for existing errors.
Sensor Vitality 2) Shows the sensor vitality.
1) Only works with the data dongle
2) Only works with the diagnostic dongle
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Communication settings
9.8 Resetting the measurement module to the factory settings
All configurable parameters are reset.
1. Select Settings > Instrument > instrument .
2. Select Confirm .
In the event of an error or prior to removing, this function can be used to deactivate dongles. Dongles are re-activated by switching the gas detector off and back on again.
1. Select Settings > Instrument > SW Dongle and the dongle to be deactivated.
2. Select Fkt. deactive .
The desired dongle is deactivated.
10.1.1
10.1.2
Setting the polling address
The polling address configures the gas detector for analogue operation or for
Multidrop operation.
HART command: #6 (Write Polling Adress)
1. Select Settings > Communication > Hart interface > Polling address .
2. Set the polling address and select Confirm .
0
1 to 15
Selects analogue operation
Selects Multidrop operation The 4-20 mA interface is deactivated and set to a constant current of approx.
3 mA.
Configure all devices of one line with different polling addresses. It is advisable to use a sequence starting with 1.
Displaying the unique identifier
The unique identifier (unique HART address) must be known for nearly all HART commands for addressing.
HART commands:
– #0 (Read Unique Identi fi er)
– #11 (Read Unique Identi fi er associated with Tag)
1. Select Settings > Communication > Hart interface > Unique identifier .
2. The unique identifier is displayed.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
10.1.3
10.2.1
10.2.2
10.2.3
10.2.4
10.2.5
Communication settings
Setting the HART tag
The HART tag (session ID designation) is used to identify a session ID.
1. Select Settings > Communication > Hart interface > Tag .
2. Set the HART tag and select Confirm . The HART tag can consist of up to 8 alphanumeric characters.
Setting the full scale value
The full scale value is equal to the measured value at which 20 mA are output. The standard full scale value of the sensor is used as default.
For information about the settings range for the full scale value refer to the instructions for use for the sensor used. For DrägerSensor O 2 and O 2 LS see
1. Select Settings > Communication > Analog interface > Analog setpoint .
2. Set the full scale value and select Confirm .
Information about the warning signal
In the event of a warning signal a warning level is issued on the 4-20-mA interface for the T2 interval. For the remainder of the time of T1, the measured signal is transmitted.
[mA]
20
4
3
T
2 t
[s]
T
1
Switching the warning signal on/off
1. Select Settings > Communication > Analog interface > Warning on/off .
2. Select Enable or Disable .
Setting the warning interval
1. Select Settings > Communication > Analog interface > Warning interval .
2. Set the times for warning intervals T1 and T2 and select bestätigen .
Setting the electric current value of the warning signal
1. Select Settings > Communication > Analog interface > Warning value .
2. Set the electric current value and select Confirm .
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Communication settings
10.2.6
10.2.7
10.2.8
10.2.9
Setting the signal type of the maintenance signal
1. Select Settings > Communication > Analog interface > Maint. signal .
2. Select the signal type.
static dynamic
A constant current signal is issued. The electric current can be configured.
An alternating signal of 3 and 5 mA with a frequency of approx. 1 Hz is issued.
[mA]
5
3
0,7 t
[s]
1,1
Adjusting the electric current of the static maintenance signal
The electric current can only be adjusted if the maintenance signal is set to static.
Otherwise the function is not available.
1. Select Settings > Communication > Analog interface > Maint. value .
2. Set the electric current value and select Confirm .
Adjusting the zero shift of the 4-20 mA signal
The 4 mA zero-point of the 4-20 mA signal can be shifted by a correction value of
± 0.5 mA.
1. Select Settings > Communication > Analog interface > Offset current .
2. Set the correction value and select Confirm .
Testing the 4-20 mA interface
For testing of the 4-20 mA interface it is possible to issue different currents as test signals simulating different states.
These functions can be used to trigger alarms in the control unit. If necessary, deactivate the alarms in the control unit first.
● To turn on a test signal, select Settings > Communication > Analog interface and the desired test.
● To turn off the test signal select the corresponding option for each test.
10.3.1 Displaying the Neuron ID
The neuron chip is the core piece of the fieldbus component. Each neuron chip has its own neuron ID by which the gas detector is recognised on the LON network.
● Select Settings > Communication > LON > Show Neuron ID .
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
10.3.2
10.4
Sensor settings
Sending the service PIN
The service PIN is a special connector of the neuron chip. The gas detector can be commissioned by sending its Neuron ID to the LON network by means of the
Service Pin function.
1. Select Settings > Communication > LON > Service-Pin .
2. Select Confirm .
The Neuron ID is sent.
Configuring the address for the Profibus PA-/
Foundation Fieldbus interface
1. Select Settings > Communication > PA/FF interface > PA Address .
2. Enter the address (settings range 001 to 126).
11.1
11.2
11.3
11.3.1
Activating the sensor replacement function
Activating this function suppresses the fault signal when the sensor is removed, and all sensor data are stored.
● Select Settings > Sensor > Sensor replacem.
.
The 4-20 mA interface issues the maintenance signal.
" Please remove sensor " is displayed.
If no entries are made, the gas detector automatically switches over to fault mode after 15 minutes.
The maintenance signal remains active until the new sensor has been installed.
Enabling/disabling the automatic calibration function
When the automatic sensor calibration function is disabled, the function cannot be selected in the menu.
1. Select Settings > Sensor > Set Autocal.
.
2. Select Enable or Disable .
Enable
Disable
Function is enabled.
Function is disabled.
Sensor test (only with sensor-test dongle or with diagnostic dongle)
Turning the sensor self-text on/off
With the sensor self-test, the gas detector continuously tests the sensor. If the sensor is not working correctly, the gas detector issues a corresponding warning or an error.
1. Select Settings > Sensor > Sensortest > Set sensortest .
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Sensor settings
11.3.2
11.4
11.5
11.6
11.7
2. Select Enable or Disable .
Performing the sensor self-test manually
1. Select Settings > Sensor and the menu item for the sensor used. Select
Sensor selftest .
2. If a sensor self-test is available (depending on the sensor used and on the device status), select Start Sensortest .
3. Select Confirm .
The result is displayed after a few seconds.
Turning the sensor lock on/off
1. Select Settings > Sensor and the menu item for the sensor used. Select
Sensorlock .
2. Select Enable or Disable .
Enable
Disable
The measurement module will accept a new sensor only if its part number, and hence the sensor type, is the same as for the old sensor.
The measurement module also accepts other sensor types and adopts the default setting of the new sensor. In this case, the configuration of the measurement module is changed.
Editing the measured gas settings
1. Select Settings > Sensor and the menu item for the sensor used. Select Gas-
Cfg .
2. First select the measured gas and then the unit of measurement.
An overview of the settings is displayed.
3. Select Confirm .
Resetting the sensor to the factory settings
1. Select Settings > Sensor and the menu item for the sensor used. Select basic init Ch .
2. Select Confirm .
The sensor is reset.
Adjusting the calibration interval
The calibration interval setting is used by the device to determine the time for the next calibration.
Calibration intervals are sensor specific. See sensor data sheet for pre-set calibration intervals and settings ranges.
1. Select Settings > Sensor and the menu item for the sensor used. Select Cal.
Intervall .
2. Set the number of days for the calibration interval.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Data storage setting (only with the data dongle)
12
12.1
12.2
12.3
12.4
12.5
12.6
3. Select Confirm .
Data storage setting (only with the data dongle)
The data logger comprises a measured value logger and an event logger.
The measured value logger can only be evaluated with the GasVision PC software
(Version 5.5 or higher).
Information about the measured value logger
The measured value logger can save at least 3000 measured values. If the sample time is set to 1 minute, the monitoring period runs to approx. 50 hours. The monitoring period can be significantly extended if the trigger mode is turned on.
Information about the event logger
The event logger stores device and sensor events (e.g. overshooting of the A1 threshold value, pump flow faults, etc.). The event logger can save 100 events.
Turning the data logger on/off
1. Select Settings > Datalogger > Datalogr. on/off .
2. Select Enable or Disable .
Setting the sample time
1. Select Settings > Datalogger > Cfg Datalogger > Sample Time .
2. Select a sample time.
Setting the evaluation mode
The evaluation mode defines whether peak values or average values are stored for the configured sample time.
1. Select Settings > Datalogger > Cfg Datalogger > Peak/Average .
2. Select an evaluation mode.
Peak
Average
Depending on the measured gas, the maximum or (for
O 2 ) minimum concentration value measured during the selected sample time is stored.
The average value of all concentration values measured during the selected sample time is stored.
Switching trigger mode on/off
With the help of the trigger mode it is possible to store measured values only from a pre-defined trigger value upward.
1. Select Settings > Datalogger > Cfg Datalogger > Trigger on/off .
2. Select an option.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Disposal
12.7
12.8
12.9
Enable
Disable
Measured values are stored only if the measured values
(with respect to the last-stored value) are greater than the configured trigger value.
All measured values are stored during the selected sample time.
Setting the trigger value
The trigger value is the threshold value from which the measured values are stored if the trigger is on. It is expressed as a percentage of the full scale value.
Example: The full scale value is 500 ppm, and a trigger value of 2 % is entered.
The data logger will then only store measured values that deviate from the last stored measured value by 10 ppm.
1. Select Settings > Datalogger > Cfg Datalogger > Trigger value .
2. Enter a value and select Confirm .
Setting the buffer mode
1. Select Settings > Datalogger > Cfg Datalogger > Stack/Roll .
2. Select an option.
Roll (Roll)
Stack (Stack)
When the storage space of the data logger is used up, the oldest data will be overwritten first.
When the storage space of the data logger is used up it will no longer be possible to store data. An appropriate warning is issued.
Deleting data from the data logger
1. Select Settings > Datalogger .
2. Select an option.
Clear Datalogger Delete measured values.
Clear Eventlogr.
Delete event data.
13 Disposal
This product must not be disposed of as household waste. This is indicated by the adjacent symbol.
You can return this product to Dräger free of charge. For information please contact the national marketing organizations or Dräger.
Measuring range and metrological properties depend on the installed sensor.
Signal transmission to the control unit
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Technical data
Analogue
1)
Measured signal
Zero-point
4 ... 20 mA
4 mA
Full-scale value 20 mA
Sensor drift below zero-point 3.8 ... 4 mA
Over range
Fault signal
Fault at the 4-20 mA interface > 23 mA
Maintenance signal Static signal: 3.4 mA (factory setting)
Range which can be set:
1.0 ... 3.5 mA (3-wire operation)
3.0 ... 3.5 mA (2-wire operation)
Dynamic alternating signal: 5 mA for
0.4 seconds and 3 mA for 0.7 seconds
Warning signal
20 ... 20.5 mA
<3.2 mA
Update rate of the 4-20-mA interface
Digital
Dynamic alternating signal between measured value and warning level.
Range which can be set (warning level):
3.0 ... 3.5 mA (factory setting: < 3.2 mA)
Intervals:
T1: 2-60 seconds
T2: 1-59 seconds and also dependent on T1
1 second
HART
®
-compatible,
Transmission via 2 or 3-core shielded cable
1) Factory setting – Depending on the configured current offset, the values can deviate by
±0,05 mA.
Overshooting or undershooting of the specified supply voltage of the gas detector can lead to faulty indication by the 4 to 20 mA signal.
For applications in compliance with Directive 2014/30/EU, the gas detector should not be operated with a direct current supply grid but with a power supply unit conforming to Protection class II or NEC Class II.
Analogue signal transmission (2-wire)
Supply voltage
(without pump or relay module) at a current of 3 mA at a current of 20 mA
AC voltage component
16.5 ... 30 V DC
8.0 V DC min.
< 0.5 V SS
Analogue signal transmission (3-wire)
Supply voltage
(without pump or relay module)
12 ... 30 V DC
AC voltage component
Load
< 0.5 V SS
0 Ohms ... 40 [Ohms/Volt] x (U S1) - 4 V)
1) Actual supply voltage on the gas detector
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Technical data
Digital signal transmission (2-wire)
Supply voltage 16.5 ... 30 V DC
AC voltage component
Load
< 0.2 V SS ; < 2.2 mV eff (500 ... 10000 Hz)
0 Ohms ... 40 [Ohms/Volt] x (U S1) - 4 V)
1) Actual supply voltage
Digital signal transmission (3-wire)
Supply voltage
(without pump or relay module)
12 ... 30 V DC
AC voltage component
Load
< 0.2 V SS
230 Ohms ... 40 [Ohms/Volt] x (U
600 Ohms max.
S1) - 4 V),
1) Actual supply voltage
Digital signal transmission (4-wire)
Supply voltage
(without pump or relay module)
12 ... 30 V DC
AC voltage component
Load
< 0.2 V SS
230 Ohms ... 40 [Ohms/Volt] x (U
600 Ohms max.
S1) - 4 V),
1) Actual supply voltage
PROFIBUS
®
PA
Communication rate
Data volume
Bus length
Segment size
Physical layer
Segment current
31.25 kBaud
244 Bytes
1900 m (6233 ft) max.
32 slaves max.
IEC 61158-2; digital, bit-synchronous, Manchester coding
18.1 mA
FOUNDATION TM Fieldbus
Communication rate
Data volume
Bus length
Segment size
Physical layer
31.25 kBaud
128 Bytes
1900 m (6233 ft) max.
240 participants max.
IEC 61158-2; digital, bit-synchronous, Manchester coding
18.1 mA Segment current
General data
Protection class (IEC 60529) IP 66, IP 67
IP 44 (if the relay module is used)
Power consumption
1)
(without analogue signal transmission)
< 50 mW
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Technical data
Cable entry
Cable core cross-section
Weight
M20 x 1.5; cable diameter 6 mm (0.24") ...
12 mm (0.47")
0.5 mm
2
(AWG 20) ... 2.5 mm
2
(AWG 13) approx. 0.9 kg (2.0 lb), without pump and relay module
4 years from delivery Life span of the buffer battery for data storage
1) To determine the overall power consumption of the gas detector in the case of use of the relay or pump module, add up the individual power consumption values.
Ambient conditions during operation
Temperature
Pressure
Humidity
–40 ... 65 °C (–40 ... 160 °F)
1)
700 ... 1300 hPa
0 ... 100 % rh, non-condensing
1) The readability of the display is impaired at temperatures below -20 °C (-5 °F) At negative temperatures, the operability of the gas detector is affected be the increasing sluggishness of the display.
Ambient conditions during storage
Temperature
Pressure
Humidity
–40 ... 70 °C (–40 ... 150 °F)
700 ... 1300 hPa
0 ... 100 % rh, non-condensing
Power supply
Operational voltage (DC)
Power consumption
12 V to 30 V at the gas detector
<2 W
Relay switch outputs logical channels principle contacts contact rating temperature resistance of the cables used
Overvoltage category
A1, A2, fault normally energised (for fail-safe operation)
1-pole changeover (SPDT)
5 A at 30 V DC; 5 A at 250 V AC at least 20 °C above the ambient temperature occurring during operation.
II
Ambient conditions
Temperature (during operation)
-40 ... 65 °C (-40 ... 160 °F)
Temperature (during storage) -40 ... 70 °C (-40 ... 150 °F)
Pressure 700 ... 1300 hPa
Humidity 0 ... 100 % rh, non-condensing
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Technical data
14.3
Supply voltage (DC) for throughput rate 0.5 L/min 12 V to 30 V at the gas detector for throughput rate 1.0 L/min 16 V to 30 V at the gas detector for throughput rate 1.5 L/min 20 V to 30 V at the gas detector
Power consumption for throughput rate 0.5 L/min <2 W for throughput rate 1.0 L/min <4 W for throughput rate 1.5 L/min <6 W
Settings
Throughput rate
Settings range approx. 0.5 L/min to 1.5 L/min (approx. 30 % to
100 %)
0.5 L/min Factory setting
Warning and alarm
Flow warning
Flow alarm
Interior diameter at the tube connection
0.4 L/min
0.3 L/min
5 mm
Ambient conditions
Temperature (during operation)
0 ... 55 °C (32 ... 130 °F)
Temperature (during storage) -40 ... 70 °C (-40 ... 150 °F)
Pressure 700 ... 1300 hPa
Humidity 0 ... 100 % rh, non-condensing
Operating conditions with DrägerSensor O 2 and O 2 LS
The DrägerSensor O 2 (6809720) is an electrochemical two-electrode sensor for the measurement of oxygen (O 2 ) in the ambient air. The sensor measures the
O 2 -partial pressure. Changes in pressure have an effect on the measured value in this process. For 1013 hPA without oxygen depletion the sensor measures
20.9 vol% O 2 .
Ambient parameters
Pressure
Humidity
Temperature
20.7 to 38.4 in. Hg (700 to 1300 hPa)
10 to 95 % rh, non-condensing
-5 to +40 °C briefly -20 to +55 °C
Storage
Pressure No effect
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Technical data
Storage
Humidity
Temperature
Duration
30 to 70 % rh, non-condensing
(only relevant if sensor packaging is open)
0 to +40 °C
Storage of sensors is not anticipated. Sensors should be used from the point of arrival. Remaining life span = expected life span - storage period
Effect of ambient parameters
Zero-point
Temperature 1)2)
-20 to 55 °C < ± 0.2 vol% O 2
Sensitivity
-10 to 55 °C
-10 to -20 °C
Pressure
Humidity
-
-
< ±0.2 vol% O
No effect
2
Relative deviation from display at 20 °C
< ± 8%
< ± 16%
Relative deviation from display at 1013 hPa:
< 10% of the measured value / 100 hPa
Relative deviation from display at 50 % rh:
< 2.5 % of the measured value
1) For temperatures below -5 °C the measurement deviation is greater than specified in EN
50104.
2) A calibration at operating temperatures must occur for operating temperature outside the range of -5 °C to 40 °C.
Reaction time
1)
0 to 25 vol% O 2 t 0...20
≤ 10 seconds t
≤
0...90
26 seconds
1) The measured value response time may increase at temperatures below -5 °C.
Stabilisation time: 5 x t 0...90
Measuring range
0-5 vol% O 2 to 0-100 vol% O 2
Standard: 25 vol% O 2
Minimum display: -1.25 vol% O 2
Warm-up time of the sensor
Operation: < 20 minutes / calibration: ≤ 2 hours
The DrägerSensor O 2 LS (6809630) is an electrochemical three-electrode sensor for the measurement of oxygen (O 2 ) in the ambient air.
Ambient parameters
Pressure 20.7 to 38.4 in. Hg (700 to 1300 hPa)
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Technical data
Ambient parameters
Humidity
Temperature
5 to 95 % rh, non-condensing
-40 to +60 °C briefly +65 °C
Storage
Pressure
Humidity
Temperature
Duration
No effect
30 to 70 % rh, non-condensing
(only relevant if sensor packaging is open)
0 to +40 °C
Storage of sensors is not anticipated. Sensors should be used from the point of arrival. Remaining life span = expected life span - storage period
Effect of ambient parameters
Zero-point
Temperature
-40 to 65 °C
Pressure
< ± 0.3 vol% O
< ± 0.1 vol% O
2
2
Sensitivity
Humidity No effect
< ± 0.3 vol% O 2
Relative deviation from display at 1013 hPa:
< 2 % of the measured value
/ 100 hPa
Relative deviation from display at 50 % rh:
< 2.5 % of the measured value
NOTICE
Cross-sensitivities
The sensor cannot be used for the measurement of oxygen if helium is present.
► Do not use the sensor for oxygen measurements if helium is present.
CAUTION
Sensor fault
If the sensor is exposed to high concentrations of unsaturated hydrocarbons, alcohol or hydrogen, this can lead to failure of the device.
► Make sure that the sensor is not exposed to such concentrations over a long period.
Reaction time
0 to 25 vol% O 2 t 0...20
≤ 10 seconds
Stabilisation time: 5 x t 0...90
t 0...90
≤ 30 seconds
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Measuring range
0-5...25 vol% O 2 , 0-25 vol% O 2
Standard: 25 vol% O 2
Minimum display: -1.25 vol% O 2
Warm-up time of the sensor
Operation: ≤ 20 minutes / calibration: ≤ 6 hours
Technical data
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Annex
15 Annex
15.1 Control drawing for UL approval 2)
2) Page 1 of the control drawing refers to the Dräger Polytron 3000 and is not pictured.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Annex
Annex
15.2 Control drawing for CSA approval
3)
ed. serv dden
rights re hten ts is forbi eac fety thority. All conten
16016 b nt /
ISO tten au
Dräger Sa tzvermerk s docume hu Sc pressed wri
Use of thi without ex
3) Page 1 of the control drawing refers to the Dräger Polytron 3000 and is not pictured.
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000 ed. serv dden
rights re hten ts is forbi eac ty. All fety conten nt /
SO 16016
authori tten k I
Dräger Sa s docume tzvermer hu pressed wri
Sc
Use of thi without ex
Annex
Annex served dden chten
rights re ts is forbi ty. All fety conten
16016 bea
authori nt / k ISO
Dräger Sa s docume
Schutzvermer ssed written
Use of thi without expre
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
15.3 Declaration of conformity
Annex
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Annex
15.4 Information about China RoHS
䜘Ԧ〠 ᴹ∂ᴹᇣ⢙䍘〠৺ਜ਼䟿 ᴹᇣ⢙䍘
䫵 (Pb) ⊎ (Hg) 䭹 (Cd) ޝԧ䬜 (Cr(VI)) ཊⓤ㚄㤟 (PBB)
兓孔◿Ⓠ䟄恾㨎 ᧤
㨎᧨6,26᧨䵾♲
P$ 㒥䘿⧉㋊兎᧨
+,&᧥ x o
㣍䯉⻞ x o ᵜ㺘Ṭᦞ SJ/T 11364 Ⲵ㿴ᇊ㕆ࡦDŽ x o o o o o
ཊⓤҼ㤟䟊
(PBDE) o o
O: 㺘⽪䈕ᴹᇣ⢙䍘൘䈕䜘Ԧᡰᴹ൷䍘ᶀᯉѝⲴਜ਼䟿൷൘ GB/T 26572 㿴ᇊⲴ䲀䟿㾱≲ԕлDŽ
X: 㺘⽪䈕ᴹᇣ⢙䍘㠣ቁ൘䈕䜘ԦⲴḀа൷䍘ᶀᯉѝⲴਜ਼䟿䎵ࠪ GB/T 26572 㿴ᇊⲴ䲀䟿㾱≲DŽ
⌘ :
10 ӗ૱ḷㆮкⲴ⧟؍֯⭘ᵏ䲀 ˄ Environmental Protection Use Period ˈ
EPUP ˅ḷ䇶㺘⽪൘↔ᵏ䰤ˈ൘↓ᑨᶑԦлˈӗ૱ѝᡰਜ਼ᴹ∂ᡆড
䲙⢙䍘ᡆᡀԭнՊਁ⭏⋴┿઼ਈᔲDŽഐ㘼↔㊫ӗ૱Ⲵ֯⭘нՊሬ㠤ԫօ ѕ䟽Ⲵ⧟ຳ⊑ḃǃԫօӪ䓛Քᇣᡆ䍒ӗᦏཡDŽ нᓄሶ↔ᵏ䰤㿶Ѫ؍؞ᵏᡆ؍䇱ᴹ᭸ᵏDŽ ḷㆮкᑖᴹ⊑ḃ᧗ࡦḷᘇⲴӗ૱ᱟਟഎ᭦Ⲵˈнᓄ䲿䘋㹼༴⨶DŽ
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
,QVWUXFWLRQVIRUXVH _ Dräger Polytron 7000
Annex
Manufacturer
Dräger Safety AG & Co. KGaA
Revalstraße 1
D-23560 Lübeck
Germany
+49 451 8 82-0
9033002 – IfU 4683.100 en
© Dräger Safety AG & Co. KGaA
Edition: 11 – 2019-09 (Edition: 1 – 2003-09)
Subject to alteration
advertisement
Related manuals
advertisement