Operating manual
Analogue refractometer
ORA 1 WB
ORA 3 WB
ORA 7 WB
ORA 2 AB
Version 1.0
12/2014
KERN ORA 1 WA/WB Wine
ORA 3 WA/WB Wine
ORA 7 WA/WB Wine
ORA 1 AB / 2 AB Alcohol
Wein
Wein
Wein
Alkohol
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: [email protected]
Tel: +49-[0]7433-9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Contents
1 technical data.................................................................................................1
2 Description......................................................................................................2
3 General information .......................................................................................3
3.1 Intended use ...................................................................................................3
3.2 Warranty..........................................................................................................3
4 Basic safety information ...............................................................................4
4.1 Follow the instructions in the operating manual .......................................4
4.2 Warning .......................................................................................................4
-5
5 supplied items ................................................................................................5
6 Before the first use ........................................................................................6
7 Use/measurement.........................................................................................6
7.1 Zero point calibration ................................................................................ 7-8
7.2 Additional advice ............................................................................................9
7.3 Measuring procedure ..............................................................................9-10
8 Cleaning and maintenance ......................................................................... 11
9 storage ......................................................................................................... 11
10 service.......................................................................................................... 12
11 Disposal ........................................................................................................ 12
12 Additional information ................................................................................ 12
13 Brix to refractive index (nD) conversion table ............................................
13
14 Annex ........................................................................................................... 14
1. technical data
Models
KeRn oRA 1 WA/ WB oRA 3 WA/ WB oRA 7 WA/ WB
ORA 1AB (Volume)
ORA 2AB (Mass)
Measuring range and Scales
Scale graduations Accuracy
Dimensions
Product
Net weight
0-140°Oe
0-25°KMW (Babo)
0-32% Brix
30-140°Oe
0-32% Brix
0-32% Brix
O R A
1 W A
0-50
O R A
3 W A
% (v/v) / (w/w)
50-80 % (v/v) / (w/w)
Messbereich und
S kalen
0 -‐1 4 0 °Oe
0 -‐2 5 °K MW
( Babo)
0 -‐3 2 % Brix
3 0 -‐1 4 0 °Oe
0 -‐3 2 % Brix
0 -‐3 2 % Brix
3 0 -‐1 4 0 °Oe
0 -‐2 5 °K MW
( Babo)
1°Oe
0,25°KMW (Babo)
0,2% Brix
1°Oe
0,2% Brix
0,2% Brix
178x40x40mm
S kalen-‐ teilung
Genauig keit
1 °Oe
0 ,2 5 °K MW
( Babo)
0 ,2 % Brix
1 °Oe
0 ,2 % Brix
1
2,5
%
%
1°Oe
178x40x40mm
Nettogewicht
(v/v) / (w/w)
0,145kg
0,145kg
0,145kg
(v/v) / (w/w)
178x40x40mm 0,145kg
178x40x40mm 0,145kg
178x40x40mm 0,145kg
198x40x40mm 0,155kg
Example scale on
ORA 1 WB
2 Gerätebeschreibung
3
4 1
7
1
1
7
2
2
2. Description
3
3
6
1. Prismenabdeckung
2. Prismenfläche
3. Justierschraube
4. Okular mit Gummi-Augenmuschel
5. Einstellring des Diopters
ORA_1WA_3WA_7WA-BA-d-1410
4
5
5 1. Prism cover
2. Prism surface
3. Adjustment screw
4. Eyepiece with rubber eyeshade
5. Diopter adjustment ring
6. Optical tubes with rubber grip
1. Prismenabdeckung
9
8
4. Okular mit Gummi-Augenmuschel
5. Einstellring des Diopters
6. Optische Röhre mit Gummigriff
10
9
8
7. Aufbewahrungsbox
8. Pipette 10
9. Refraktometer
10. Destilliertes Wasser / Kalibrierflüssigkeit
11. Justierwerkzeug
12. Reinigungstuch
12
12
11
7. Aufbewahrungsbox
11
7. Storage box
8. Pipette
9. Refractometer
10. C alibration liquid
(Distilled water)
11. Adjustment tool
12. Cleaning cloth
4
10. Destilliertes Wasser / Kalibrierflüssigkeit
11. Justierwerkzeug
12. Reinigungstuch
2
ORA_1WB_3WB_7WB_2AB-BA-d-1410 4
3. General information
3.1 Intended use
The refractometer is a measuring instrument for determining the refractive index of transparent substances in the liquid or the solid state. It is used to observe the behaviour of light as it passes from a prism with known properties to the substance being tested.
Use of the refractometer for other purposes is contrary to its intended use and may be hazardous. The manufacturer shall not be liable for any damages caused by improper use.
3.2 Warranty
The warranty shall be void in the event of:
▸ Failure to observe the instructions in the operating manual
▸ Use for purposes other than those described
▸ Modifications or opening the device housing
▸ Mechanical damage and/or damage resulting from media, liquids, natural wear and tear
3
4. Basic safety information
4.1 Follow the instructions in the operating manual
▸ Carefully read through the operating manual even if you have prior experience with KERN refractometers.
▸ Every language version includes a non-authoritative translation.
The original German document is the definitive version.
4.2 Warning
▸ Do not let acids come into contact with skin or eyes. If acid comes into contact with skin, flush with copious amounts of water. Shower if larger areas of skin are affected.
▸ If acid comes into contact with eyes, keep the eyelid open and flush the eye with running lukewarm water from the outer corner to the inner corner. Flush eyes for at least 15 minutes. Then consult a doctor or ophthalmologist immediately.
▸ Thoroughly clean the refractometer after each use.
▸ The refractometer must not be exposed to extreme temperatures, high mechanical stresses, strong direct sunlight or high humidity.
▸ This refractometer is not a toy. Keep out of reach of children.
▸ Make sure that you will not be hit by anything else while you are using the refractometer, as this could cause serious eye injuries
▸ The rubber eyeshade may cause irritation when in prolonged contact with the skin. If this happens, consult your doctor.
▸ Do not touch the lenses with your fingers.
4
5. supplied items
After unpacking and before using the device for the first time, check that all listed parts have been supplied. Replace damaged or faulty parts immediately and do not put them into operation.
▸ Refractometer
▸ Storage box
▸ Pipette
▸ Adjustment tool
▸ Cleaning cloth
▸ C alibration liquid (Distilled water)
6. Before the first use
Remove the protective film (if present) from the prism surface [2] and check that the rubber eye-cup [4] is fitted correctly.
7. Use/measurement
The refractometer can be used to quickly and accurately determine the refractive index of transparent substances, liquid or solid ones. To ensure correct measurement, the measuring device should be adjusted before measurements are carried out. Please make sure your hands are dry before handling the measuring device.
5
6
7.1 Zero point calibration
▸ Hold up the refractometer against a sufficiently bright light source and look through the eyepiece [5], holding the rubber eyeshade [4] closely against your eye/glasses.
▸ Rotate the eyepiece [5] to adjust it for your eyesight, until you can see the scale sharply.
▸ Open the prism cover [1].
▸ Thoroughly clean the prism [2] and the underside of the prism cover [1] using a soft cloth or soft paper (with alcohol if necessary) and wipe dry.
▸ Now apply a few drops of the calibration liquid [10] onto the prism surface [2].
▸ C lose the prism cover [1]. The amount of fluid should be sufficient to moisten most of the prism surface. There should not be any air bubbles between the measuring prism [2] and prism cover [1].
▸ Wait about 30 seconds to allow the temperatures of the fluid and prism to equalise.
▸ Look through the eyepiece [4] while pointing the refractometer‘s prism surface [2] at a bright light source.
▸ Through the eyepiece [4], you will see a bright and a blue field. The boundary line between them shows the measured value on the scale which is also visible through the eyepiece [4].
7
▸ Use the supplied adjustment tool [11] to turn the adjustment screw [3] behind the prism surface [2] (under the rubber cap), and adjust the scale so that the measuring device is optimally set up by moving the boundary line upwards or downwards.
Calibration value ORA 1WA/1WB: 0 % (Brix) / 0 °Oe (Oechsle) / 0 °KMW (KMW Babo)
Calibration value ORA 3WA/3WB: 0 % (Brix)
Calibration value ORA 7WA/7WB: 0 % (Brix) / 0 °Oe (Oechsle) / 0 °KMW (KMW Babo)
Calibration value ORA 1AB / 2AB: 0 % (v/v) / (w/w)
▸ Repeat step 4 (cleaning).
Important!
The ambient/room temperature and the sample temperature influ-ence the refractometer measuring result. The scales of the refractometer models,
!
which have " W B / AB " in its name, are designed for an ambient temperature of +20 °C! If the measurements are carried out at a temperature other than +20 °C, the results must be correspondingly corrected. A correction table can be found in the annex,
Point 14. Refractometer models, which have " W A" in its name, are equipped with an automatic temperature compensation (ATC). Between +10°C and +30°C measurement differences due to temperature changes are automatically compensated.
7.2 Additional advice
It is important that the samples being measured are representative samples. For samples that evaporate easily, measurements should be carried out quickly. The samples should be at the same temperature as the measuring instrument in order to achieve an accurate result.
7.3 Measuring procedure
▸ Make sure your hands are dry before handling the refractometer.
▸ Open the prism cover [1] and use the supplied pipette [8] to apply a few drops of the sample liquid [8] onto the prism surface [2]. Close the prism cover [1]. Spread the liquid evenly by pressing down on the prism cover [1] and elimi-nate any air bubbles present.⇨
▸ Hold the device horizontally and wait about 30 seconds (for optimal tempera-ture equalisation between the sample and device).
▸ View the measurement scale through the eyepiece [4]. Point the prism surface [2] of the refractometer at a bright light source while doing this.
▸ Rotate the adjustment ring [5] on the eyepiece [4] to adjust the focus.
▸ The boundary line will move on the measurement scale depending on the concentration. This bright/dark boundary line shows the result on the scale.
▸ If the temperature deviates from +20 °C and a refractometer without ATC is used , correct the measured result using the corresponding value from the temperature correction table [13].
▸ Carefully clean the supplied pipette [8] and the refractometer after carrying out the measurement.
8
Important!
After every measurement, use a lint-free, absorbent cloth to remove the fluids from the prism surface [2] and prism cover [1]. Then carefully clean the prism and prism cover using a cloth moistened with water or if necessary alcohol, and dry both parts using a soft, dry and lint-free cloth. Avoid rubbing the prism [2].
!
10
8. Cleaning and maintenance
Clean the refractometer using a soft, lint-free cloth moistened with either water or, if necessary, alcohol. Do not use any aggressive or abrasive cleaning agents.
Never immerse the device in water or hold it under running water. Never handle the device with wet or damp hands.
Never touch the measuring prism [2] with hard tools made from plastic, wood, rubber, metal, glass etc. Hard objects can quickly damage the relatively soft prism glass, resulting in measurement errors.
The refractometer is maintenance-free.
Cleaning should be carried out immediately before and after each use of the refractometer to maximise its life and optimise measurement results.
9. storage
Store the refractometer in a dry, non-corrosive environment, preferably between 10 °C and 30 °C.
10. service
After reading this operating manual, if you have any questions about setting up or using the refractometer, or if any unexpected problem occurs, please contact your dealer. The device housing may only be opened by trained service technicians authorised by KERN.
11
11. Disposal
The packaging consists of environmentally friendly materials which can be disposed of via local recycling facilities.
The device and storage box should be disposed of by the operator in accordance with applicable national or regional regulations at the place of use.
12. Additional information
The product may differ slightly from the illustrations. Avoid exposing the refractometer to direct sunlight. Never bring the refractometer into contact with solvents.
13. Brix to refractive index (nD) conversion table
Data from „ICUMSA“ International Commission for Uniform Methods of Sugar Analysis, at 20 °C and 589 nm wavelength.
12
13
14. Annex
Table 1: International Temperature Correction Table for °Brix (% sugar gradient)
Correct the result by the following values (refractometer must be correctly adjusted at 20 °C).
18.0
19.0
20.0
21.0
22.0
23.0
24.0
25.0
10.0
11.0
12.0
13.0
14.0
15.0
16.0
17.0
26.0
27.0
28.0
29.0
30.0
% Brix reading
0.0 5.0 10.0 15.0 20.0 25.0 30.0 35.0 40.0 45.0 50.0 55.0 60.0 65.0 70.0 75.0 80.0 85.0
-0.53 -0.56 -0.59 -0.62 -0.65 -0.67 -0.69 -0.71 -0.72 -0.73 -0.74 -0.75 -0.75 -0.75 -0.75 -0.75 -0.74 -0.73
-0.49 -0.52 -0.54 -0.57 -0.59 -0.61 -0.63 -0.64 -0.65 -0.66 -0.67 -0.68 -0.68 -0.68 -0.68 -0.67 -0.67 -0.66
-0.44 -0.47 -0.49 -0.51 -0.53 -0.55 -0.56 -0.57 -0.58 -0.59 -0.60 -0.60 -0.61 -0.61 -0.60 -0.60 -0.60 -0.59
-0.40 -0.41 -0.43 -0.45 -0.47 -0.48 -0.50 -0.51 -0.52 -0.52 -0.53 -0.53 -0.53 -0.53 -0.53 -0.53 -0.52 -0.52
-0.34 -0.36 -0.38 -0.39 -0.40 -0.42 -0.43 -0.44 -0.44 -0.45 -0.45 -0.46 -0.46 -0.46 -0.46 -0.45 -0.45 -0.44
-0.29 -0.31 -0.32 -0.33 -0.34 -0.35 -0.36 -0.37 -0.37 -0.38 -0.38 -0.38 -0.38 -0.38 -0.38 -0.38 -0.37 -0.37
-0.24 -0.25 -0.26 -0.27 -0.28 -0.28 -0.29 -0.30 -0.30 -0.30 -0.31 -0.31 -0.31 -0.31 -0.31 -0.30 -0.30 -0.30
-0.18 -0.19 -0.20 -0.20 -0.21 -0.21 -0.22 -0.22 -0.23 -0.23 -0.23 -0.23 -0.23 -0.23 -0.23 -0.23 -0.23 -0.22
-0.12 -0.13 -0.13 -0.14 -0.14 -0.14 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15
-0.06 -0.06 -0.07 -0.07 -0.07 -0.07 -0.07 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.07
0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00
0.06 0.07 0.07 0.07 0.07 0.07 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.07
0.13 0.14 0.14 0.14 0.15 0.15 0.15 0.15 0.16 0.16 0.16 0.16 0.16 0.16 0.15 0.15 0.15 0.15
0.20 0.21 0.21 0.22 0.22 0.23 0.23 0.23 0.23 0.24 0.24 0.24 0.24 0.23 0.23 0.23 0.23 0.22
0.27 0.28 0.29 0.29 0.30 0.30 0.31 0.31 0.31 0.32 0.32 0.32 0.32 0.31 0.31 0.31 0.30 0.30
0.34 0.35 0.36 0.37 0.38 0.38 0.39 0.39 0.40 0.40 0.40 0.40 0.40 0.39 0.39 0.39 0.38 0.37
0.42 0.43 0.44 0.45 0.46 0.46 0.47 0.47 0.48 0.48 0.48 0.48 0.48 0.47 0.47 0.46 0.46 0.46
0.50 0.51 0.52 0.53 0.54 0.55 0.55 0.56 0.56 0.56 0.56 0.56 0.56 0.55 0.55 0.54 0.53 0.52
0.58 0.59 0.60 0.61 0.62 0.63 0.64 0.64 0.64 0.65 0.65 0.64 0.64 0.64 0.63 0.62 0.61 0.60
0.66 0.67 0.68 0.69 0.70 0.71 0.72 0.73 0.73 0.73 0.73 0.73 0.72 0.72 0.71 0.70 0.69 0.68
0.74 0.75 0.77 0.78 0.79 0.80 0.81 0.81 0.81 0.82 0.81 0.81 0.81 0.80 0.79 0.78 0.77 0.75
14
Data from "ICUMSA" International Commission for Uniform Methods of Sugar Analysis, at 20
ORA_1SB_2SB_3SB-BA-e-1510