SWIFT SW350B is a versatile microscope suitable for both beginners and professionals in various fields. With its 40X, 100X, 250X, 400X, 1000X, and 2500X magnifications, it allows for detailed observation of microscopic specimens. The built-in 1W LED illumination ensures clear and bright images, while the adjustable iris diaphragm and condenser provide precise control over the amount of light. The 360° rotatable binocular head and wide-field eyepieces offer a comfortable viewing experience.
Advertisement
Advertisement
User Guide: SW380B
Welcome to your Swift SW380B microscope! Whether you are a seasoned professional or a beginner hobbyist, Swift has the perfect microscope for you.
This quick-start guide will introduce you to the components of your new microscope and help you begin to use it. For more information about our microscopes or customer service requests, please visit www.swiftoptical.com.
Table of Contents
VII. Warranty information and customer support
1
I. Specifications
Model
Head
SW380B
360° rotatable Binocular Siedentopf
Inclination
Eyepieces
30°
Wide-field 10X and 25X pairs
Interpupillary distance 48-75mm
Diopter adjustment On left tube, +/-5dp
Nosepiece
Objectives
Quadruple revolving
Achromatic
4X, NA 0.1, working distance 37.5mm
10X, NA 0.25, working distance 6.54mm
40X(Spring), NA 0.65, working distance 0.63mm
100X (Spring, Oil), NA 1.25, working distance 0.195mm
Magnifications
Stage
Stage size
Stage X-Y range
40X, 100X, 250X, 400X, 1000X and 2500X
Double-layered mechanical stage with slide holder
130mmx130mm
70mmx30mm
Upper limit stop
Condenser
Rack stop preset but adjustable
NA 1.25 with iris diaphragm
Focusing Coaxial coarse and ultra-fine focusing system
Fine focus precision 0.002mm increments
Illumination
Power
Built-in 1W LED transmitted illumination with intensity control
110-240V, built-in transformer
2
II. Parts of the microscope
III. Assembly and set up
Carefully remove the microscope body from the box and set it on a sturdy, flat surface.
Remove the microscope head from the box and secure it to the body with the head locking screw, using care not to overtighten.
Remove the plastic covers from the eyepiece holders on the microscope's head, then place eyepieces into the holders. Locate the power cord and attach it to the microscope, then plug it in. Turn the microscope on by pressing the power button on the back of the base.
IV. Using the microscope
Turn the microscope on and use the illumination intensity control wheel to set the lamp on the lowest brightness setting.
3
Clip a slide into the slide holder on the stage. Use the X-Y translational control knobs on the right side of the stage to move the stage front to back or the slide holder left to right to center the specimen over the bulb.
Move the binocular eyepieces closer together or further apart until you see one image when you look through them. Adjust the bulb brightness to a comfortable intensity with the dimmer wheel.
Always start with the lowest power objective lens, in this case the 4X objective, centered over the stage. Look through the eyepieces and bring the image into focus by slowly turning the larger coarse adjustment focus knob to bring the stage closer to the objective. Once an image forms, switch to the blue fine adjustment focus knob to fine-tune for clarity or to see the various layers of the specimen you are examining. You may also use the diopter adjustment control on the eyepiece holder at this time to further fine-tune image clarity. Look at other parts of the slide by using the X-Y translational control knobs to move the slide holder once the specimen is in focus.
Increase magnification by rotating the nosepiece clockwise to a higher magnification objective, taking care not to strike the slide with the objective as doing so may damage the glass on the slide, the objective, or both. If you cannot move the objective into place without making contact with the slide, use the coarse focus knob to move the stage down and away to make room before changing magnification.
Adjust the focusing ring on the iris aperture diaphragm to change the resolution of the image. The condenser can be manually moved closer to or farther from the stage by rotating the metal knob.
For best results with the 100X objective, use the included immersion oil. Rotate the nosepiece so that the 100X objective is near, but not on the specimen slide. Then place a drop of the included immersion oil onto the cover slip. Slowly rotate the 100X objective into place so that it comes into contact with the immersion oil. Adjust focus until your specimen comes clearly into view. (Note: Immediately after use, always gently wipe the oil off of the objective lens and the cover slip with lens paper moistened with a small amount of ethyl or isopropyl alcohol. Do not use immersion oil with the 4X, 10X, or 40X objectives.)
4
V. Caring for your microscope
As with any quality instrument, your microscope should be stored in a cool, dry place.
Use the dust cover to protect the microscope when not in use. If necessary, clean the lenses with a soft brush or compressed air to gently remove dust.
Avoid touching the optical surfaces directly and never disassemble the optical components of the microscope. When not using eyepieces, use the plastic covers to protect the eyepiece lens holders and ocular lens.
VI. Troubleshooting
If you have a problem, you may be able to correct it yourself. Here are some common issues and easy solutions you can try before calling customer support for service.
*Caution: Never disassemble the electrical, mechanical, or optical components. This servicing should only be done by a Swift technician.
Issue Possible cause Solution
Visual field is uneven, or image is dark at the edges
Nosepiece is not locked in place to center the objective lens
Turn nosepiece until you hear a click to center the objective lens
Slide is not centered over the bulb, leaving part of the specimen unilluminated
Center the slide over the bulb with the X-Y translational control knobs
Dirty objective or eyepiece lens
Iris aperture is closed too far
Clean with compressed air or soft brush
Open the iris aperture wider
Visual field is blurry or dirty Dirty objective or eyepiece lens
The slide is dirty
Clean with compressed air or soft brush
Clean the slide with glass cleaner
Eyepiece or objective lens not fully screwed into place
Tighten eyepiece or objective lenses
5
Poor image quality No cover slip on the specimen
Place glass cover slip over the specimen
Slide is too thick or too thin Use a lab-quality slide
Slide is placed upside down on the stage
Oil on objective lens
Flip the slide over so that the cover slip is facing up
Wipe the lens off with soft cloth and glass cleaner
Adjust the iris aperture Iris aperture is too narrow or too wide
Slide is shifting on the stage
Reposition slide more securely in the slide holder
Image moves while adjusting focus
Light is too dim
Image won't come into focus when using a higher-powered objective
Iris aperture is closed too far
Bulb setting is too low
Condenser lens is dirty
Slide is upside down
Open the iris aperture wider
Increase intensity on the control wheel
Wipe the condenser lens off with a soft cloth
Voltage is too low Use the supplied power adapter
Objective lens collides with the slide when switching from a lower to a higher power objective
Slide is upside down or too thick
Use a lab-quality slide or flip the slide over so the cover slip is facing up
Stage is set too close to the objectives
Safety rack stop needs adjustment
Lower the stage with the coarse focus knob
Adjust the safety rack stop to stop the stage farther from the objectives
Flip the slide over so the cover slip is facing up
6
Slide is not moving when using the X-Y translational control knob
Lamp does not turn on
Slide has come unclipped from the holder
No power source
Lamp flickers while on
Reposition slide in the holder and secure with clip
Check that the adapter is fully connected or that batteries are installed
Bulb has burned out Replace the bulb
Bulb is not inserted into the socket correctly
Check to make sure the bulb is fully screwed into the socket
Power adapter is not fully plugged in
Check to make sure that the adapter is fully connected to wall socket and to the microscope
VII. Warranty information and customer support
Our microscopes are manufactured to meet ISO 9001 standards. Swift warranties are as follows:
● Five (5) Year Warranty for Microscopes: Microscopes come with a five (5) year warranty against manufacturing defects. Does not cover normal wear, routine maintenance, add-on accessories, damage resulting from repair by unauthorized parties, accident, alteration, shipping, misuse or abuse is not covered.
● One (1) Year Warranty for Electrical and Video components. Does not cover light bulbs, batteries, fuses, or electrical cords.
All warranties start from the original date of purchase. Swift provides the repair or replacement of warrantied parts for free, including labor, during the warranty period.
Proof of original purchase is required. Buyers are responsible for shipping to and from our warehouse for warranty services. The warranty does not cover damages resulting from normal wear and tear, abuse, or unauthorized repairs. Warranty service is provided by Swift Optical Instruments, Inc.’s authorized technicians. Determination of warranty is at the technician’s discretion.
7
*For customers living outside the United States, Swift Optical Instruments, Inc. will provide standard warranty service. Both inbound and outbound shipping costs
(including duties and taxes) is the responsibility of the consumer.
For more information or to submit a repair request, please contact our Customer
Support department:
Tel: 877-967-9438, option 1
Email: [email protected]
Disclaimer: We are constantly working to improve our instruments and to adapt them in response to customer feedback. These improvements occasionally involve small modifications to the mechanical structure and optical design of our microscopes.
Therefore, some descriptions, illustrations, and specifications in this instruction manual may vary slightly from the microscope you receive.
Swift Optical Instruments, Inc.
(877) 967-9438 www.swiftoptical.com
8