DK32 - DK34 - DK37

Handbook

Variable area flowmeters

Electronic revision ESK3x: ER 1.1.x

© KROHNE 09/2020 - 4000287107 - MA DK32-34-37 R09 en

:

IMPRINT

:::::::::::::::::::::::::::::::::::::::

All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtechnik GmbH.

Subject to change without notice.

2

Copyright 2020 by

KROHNE Messtechnik GmbH - Ludwig-Krohne-Str. 5 - 47058 Duisburg (Germany) www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

CONTENTS

1 Safety instructions 5

1.1 Electronic revision DK32/ESK, DK34/ESK ....................................................................... 5

1.2 Intended use ..................................................................................................................... 5

1.3 Certifications .................................................................................................................... 6

1.4 Pressure equipment directive.......................................................................................... 7

1.5 Safety instructions from the manufacturer ..................................................................... 8

1.5.1 Copyright and data protection ................................................................................................ 8

1.5.2 Disclaimer ............................................................................................................................... 8

1.5.3 Product liability and warranty ................................................................................................ 9

1.5.4 Information concerning the documentation........................................................................... 9

1.5.5 Warnings and symbols used................................................................................................. 10

1.6 Safety instructions for the operator............................................................................... 10

2 Device description 11

2.1 Scope of delivery............................................................................................................. 11

2.2 Device versions............................................................................................................... 12

2.3 Nameplate ...................................................................................................................... 14

2.4 Description code............................................................................................................. 15

3 Installation 16

3.1 General notes on installation ......................................................................................... 16

3.2 Storage ........................................................................................................................... 16

3.3 Transport ........................................................................................................................ 16

3.4 Installation conditions .................................................................................................... 16

4 Electrical connections 17

4.1 Safety instructions.......................................................................................................... 17

4.2 Electrical connection of limit switches .......................................................................... 18

4.2.1 Electrical connection DK32 / DK34....................................................................................... 18

4.2.2 Electrical connection DK37 M8M.......................................................................................... 19

4.2.3 Settings of NAMUR limit switches for DK32, DK34, DK37 M8M .......................................... 20

4.2.4 Settings of Reed limit switch ................................................................................................ 20

4.3 Electrical signal output ESK3x for DK32, DK34............................................................. 21

4.3.1 Power supply......................................................................................................................... 22

4.3.2 Load for HART communication............................................................................................. 22

4.3.3 Configuration......................................................................................................................... 22

4.4 Electrical signal output for DK37/M8E........................................................................... 23

4.4.1 Power supply......................................................................................................................... 24

4.4.2 Load for HART communication............................................................................................. 24

4.4.3 Configuration......................................................................................................................... 24

4.5 Grounding connections................................................................................................... 25

4.6 Ingress protection .......................................................................................................... 26

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

3

4

CONTENTS

DK32 - DK34 - DK37

5 Start-up 27

5.1 Start-up........................................................................................................................... 27

6 Service 28

6.1 Maintenance ................................................................................................................... 28

6.2 Spare parts availability................................................................................................... 29

6.3 Availability of services .................................................................................................... 29

6.4 Returning the device to the manufacturer..................................................................... 29

6.4.1 General information.............................................................................................................. 29

6.4.2 Form (for copying) to accompany a returned device............................................................ 30

6.5 Disposal .......................................................................................................................... 30

6.6 Disassembly and recycling............................................................................................. 31

6.6.1 Description of the device components ................................................................................. 31

6.6.2 Indicator versions ................................................................................................................. 32

7 Technical data 34

7.1 Functional principle........................................................................................................ 34

7.2 Technical data................................................................................................................. 35

7.2.1 Temperatures for all device versions................................................................................... 37

7.2.2 Electrical connection data for indicator DK32, DK34, DK37/M8M with limit switch ........... 38

7.2.3 Electrical connection data for transmitter ESK3x for DK32, DK34 ..................................... 39

7.2.4 Electrical connection data for indicator DK37/M8E with electrical signal output .............. 40

7.2.5 Approvals .............................................................................................................................. 41

7.3 Dimensions and weight .................................................................................................. 42

7.3.1 DK32, DK34 ........................................................................................................................... 42

7.3.2 DK32, DK34 with transmitter housing ESK3x ...................................................................... 44

7.3.3 DK37/M8M............................................................................................................................. 45

7.3.4 DK37/M8E.............................................................................................................................. 46

7.3.5 Minimum distances when installing several measuring devices ........................................ 47

7.4 Measuring ranges........................................................................................................... 48

7.4.1 Float ...................................................................................................................................... 48

7.4.2 Valves (only DK32 and DK37) ................................................................................................ 49

7.5 Flow regulators for variable pressure (only DK32 and DK37)....................................... 50

7.5.1 Control ranges ...................................................................................................................... 51

7.5.2 Technical data for flow regulators ....................................................................................... 52

7.5.3 Dimensions with flow regulator ........................................................................................... 53

8 Notes 54 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

SAFETY INSTRUCTIONS

1

1.1 Electronic revision DK32/ESK, DK34/ESK

The electronic revision of the ESK3x indicates the respective hardware/software status of the electronics.

Electronic revision Explanation

ER 1.1.x

Basic version

Table 1-1: Description of the electronic revision

Documentation

MA DK32-37 R07, R08, R09

1.2 Intended use

CAUTION!

Responsibility for the use of the measuring devices with regard to suitability, intended use and corrosion resistance of the used materials against the measured fluid lies solely with the operator.

INFORMATION!

This device is a Group 1, Class A device as specified within CISPR11:2009. It is intended for use in industrial environment. There may be potential difficulties in ensuring electromagnetic compatibility in other environments, due to conducted as well as radiated disturbances.

INFORMATION!

The manufacturer is not liable for any damage resulting from improper use or use for other than the intended purpose.

The variable area flowmeters are suitable for measuring gases and liquids.

INFORMATION!

• The product may not contain any ferromagnetic particles or solids.

The product must be sufficiently liquid and free of deposits.

Avoid pressure surges and pulsing flows.

Open valves slowly.

Do not use solenoid valves.

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

5

1

SAFETY INSTRUCTIONS

DK32 - DK34 - DK37

The devices are particularly suitable for the measurement of small quantities of:

• Process or carrier gases

• Nitrogen, CO

2

or other industrial gases

• Sample flows for process analysers

• Sealing gas or sealing liquid measurement on sealing systems

• Purge fluids for measuring systems

• Air or water

• Chemicals and additives

• Lubricating, cooling and anti-corrosive agents

DANGER!

For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.

CAUTION!

Do not use any abrasive media containing solid particles.

1.3 Certifications

CE marking

The device fulfils all applicable statutory requirements of the EU directives:

• Pressure equipment directive

• For devices with electrical installations: EMC directive

• Devices for use in hazardous areas: ATEX directive

The manufacturer certifies successful testing of the product by applying the CE mark.

An EU declaration of conformity regarding the directives in question and the associated harmonised standards can be downloaded from our website.

6 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

SAFETY INSTRUCTIONS

1

1.4 Pressure equipment directive

A conformity assessment in accordance with pressure equipment directive has been carried out for the devices described.

The PED key describes the rating of the devices:

Example: PED/G1/4.3/SEP

G

1

4.3

SEP

Gases and vapours

Fluid group 1

Article 4.3 of the directive 2014/68/EU

Sound engineering practice

Table 1-2: Example of PED key

The PED key identification can be found on the nameplate of the device (for details refer to

Nameplate on page 14).

INFORMATION!

The stated pressures (PS) and temperatures (TS) only apply as refers to the pressure resistance of the sensor body. As regards the functionality of the entire device, further restrictions of the maximum temperature may need to be observed (e.g. ATEX approval). Devices rated below category I due to their size, do not receive the CE mark in the scope of the PED. These devices are subject to applicable sound engineering practice (SEP).

Residual risk

A risk analysis in accordance with the pressure equipment directive has been carried out for the devices. The residual risk is described as follows:

• The devices are designed according to the valid and applicable rules and standards for static operation and their pressure resistance is calculated for the declared maximum pressure and temperature (no calculation for cyclical change).

• Responsibility for the use of the measuring devices with regard to corrosion resistance of the used materials against the measured fluid lies solely with the operator.

• Avoid abrasion.

• Avoid pulsation and cavitation.

• Protect devices from vibration and high-frequency oscillation.

• Draining (backflow) may be delayed due to the float in the measuring tube. Expect residue around the valve and in the flow regulator.

• Implement appropriate measures to counteract external fire hazards.

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

7

1

SAFETY INSTRUCTIONS

DK32 - DK34 - DK37

1.5 Safety instructions from the manufacturer

1.5.1 Copyright and data protection

The contents of this document have been created with great care. Nevertheless, we provide no guarantee that the contents are correct, complete or up-to-date.

The contents and works in this document are subject to copyright. Contributions from third parties are identified as such. Reproduction, processing, dissemination and any type of use beyond what is permitted under copyright requires written authorisation from the respective author and/or the manufacturer.

The manufacturer tries always to observe the copyrights of others, and to draw on works created in-house or works in the public domain.

The collection of personal data (such as names, street addresses or e-mail addresses) in the manufacturer's documents is always on a voluntary basis whenever possible. Whenever feasible, it is always possible to make use of the offerings and services without providing any personal data.

We draw your attention to the fact that data transmission over the Internet (e.g. when communicating by e-mail) may involve gaps in security. It is not possible to protect such data completely against access by third parties.

We hereby expressly prohibit the use of the contact data published as part of our duty to publish an imprint for the purpose of sending us any advertising or informational materials that we have not expressly requested.

1.5.2 Disclaimer

The manufacturer will not be liable for any damage of any kind by using its product, including, but not limited to direct, indirect or incidental and consequential damages.

This disclaimer does not apply in case the manufacturer has acted on purpose or with gross negligence. In the event any applicable law does not allow such limitations on implied warranties or the exclusion of limitation of certain damages, you may, if such law applies to you, not be subject to some or all of the above disclaimer, exclusions or limitations.

Any product purchased from the manufacturer is warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale.

The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes.

8 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

SAFETY INSTRUCTIONS

1

1.5.3 Product liability and warranty

The operator shall bear responsibility for the suitability of the device for the specific purpose.

The manufacturer accepts no liability for the consequences of misuse by the operator. Improper installation or operation of the devices (systems) will cause the warranty to be void. The respective "Standard Terms and Conditions" which form the basis for the sales contract shall also apply.

1.5.4 Information concerning the documentation

To prevent any injury to the user or damage to the device it is essential that you read the information in this document and observe applicable national standards, safety requirements and accident prevention regulations.

If this document is not in your native language and if you have any problems understanding the text, we advise you to contact your local office for assistance. The manufacturer can not accept responsibility for any damage or injury caused by misunderstanding of the information in this document.

This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of icons as shown below.

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

9

1

SAFETY INSTRUCTIONS

DK32 - DK34 - DK37

1.5.5 Warnings and symbols used

Safety warnings are indicated by the following symbols.

DANGER!

This warning refers to the immediate danger when working with electricity.

DANGER!

This warning refers to the immediate danger of burns caused by heat or hot surfaces.

DANGER!

This warning refers to the immediate danger when using this device in a hazardous atmosphere.

DANGER!

These warnings must be observed without fail. Even partial disregard of this warning can lead to serious health problems and even death. There is also the risk of seriously damaging the device or parts of the operator's plant.

WARNING!

Disregarding this safety warning, even if only in part, poses the risk of serious health problems.

There is also the risk of damaging the device or parts of the operator's plant.

CAUTION!

Disregarding these instructions can result in damage to the device or to parts of the operator's plant.

INFORMATION!

These instructions contain important information for the handling of the device.

LEGAL NOTICE!

This note contains information on statutory directives and standards.

• HANDLING

This symbol designates all instructions for actions to be carried out by the operator in the specified sequence.

i RESULT

This symbol refers to all important consequences of the previous actions.

1.6 Safety instructions for the operator

WARNING!

In general, devices from the manufacturer may only be installed, commissioned, operated and maintained by properly trained and authorized personnel.

This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device.

10 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

DEVICE DESCRIPTION

2

2.1 Scope of delivery

INFORMATION!

Inspect the packaging carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.

INFORMATION!

Do a check of the packing list to make sure that you have all the elements given in the order.

INFORMATION!

Look at the device nameplate to ensure that the device is delivered according to your order.

Check for the correct supply voltage printed on the nameplate.

Figure 2-1: Scope of delivery

1 Measuring device in ordered version

2 Product documentation

3 Certificates, calibration report (supplied to order only)

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

11

2

DEVICE DESCRIPTION

2.2 Device versions

• DK32 with valve and horizontal connection

• DK34 without valve and vertical connection

• DK37 with valve and horizontal connection

• DK37 without valve and vertical connection

The following designs are available as options:

• with flow regulator for variable inlet pressure

• with flow regulator for variable outlet pressure

• with flange adapter (total length: 250 mm / 9.8")

• with indicator housing DK37 in stainless steel

• with limit switches or 4...20 mA/HART signal output

DK32 - DK34 - DK37

Figure 2-2: Device versions

1 DK32 with valve and horizontal connection

2 DK34 without valve and vertical connection

3 DK37/M8E with valve and electronic indicator

4 DK37/M8M with valve and mechanical indicator

12

Figure 2-3: Versions with regulators

1 DK32 with flow regulator for variable inlet pressure

2 DK32 with flow regulator for variable outlet pressure www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

DEVICE DESCRIPTION

2

Figure 2-4: Versions with flange adapter

1 DK34 with DN15/PN40 adapter

2 DK32 with DN15/PN40 adapter

Figure 2-5: Version DK37 with stainless steel housing

1 DK37/M8E/R

2 DK37/M8M/R

Figure 2-6: Versions DK32 with limit switches or 4...20 mA/HART signal output

1 DK32 with NAMUR limit switches

2 DK32 with 4...20 mA/HART signal output

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

13

2

DEVICE DESCRIPTION

DK32 - DK34 - DK37

2.3 Nameplate

INFORMATION!

Look at the device nameplate to ensure that the device is delivered according to your order.

Check for the correct supply voltage printed on the nameplate.

Figure 2-7: Examples of nameplates

1 Device type

2 Manufacturer

3 Identification number of the notified body

4 Rating data: temperature & pressure rating

5 Data according to the pressure equipment directive

6 Marking according to the notified body

7 Electrical connection data

8 Data Matrix Code

9 Note to observe the documentation and for disposal

Additional markings on the measuring device:

• SO - sales order / item

• PA - production order

• Vx - product configurator code

• AC - article code

14 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

DEVICE DESCRIPTION

2

2.4 Description code

The description code consists of the following elements *:

Figure 2-8: Description code DK32, DK34

1 32 - with valve and horizontal connection

34 - without valve and vertical connection

2 RE - flow regulator for variable inlet pressure

RA - flow regulator for variable outlet pressure

3 K1 - one limit switch, type NAMUR

K2 - two limit switches, type NAMUR

R1 - one limit switch, type Reed

ESK - 4...20 mA/HART ®

4 S - plug connector

L - cable entry including cable

5 HT - high-temperature version

6 free

7 Ex - explosion-protected equipment

8 SK - SIL conformity according to IEC 61508 of the limit switch

* positions which are not needed are omitted (no blank positions)

Figure 2-9: Description code DK37

1 M8M - mechanical indicator

M8E - electronic indicator

2 without - indicator housing in PPS

R - indicator housing in stainless steel

3 RE - flow regulator for variable inlet pressure

RA - flow regulator for variable outlet pressure

4 K1 - one limit switch

K2 - two limit switches

5 Ex is not part of the description code

6 SK - SIL conformity according to IEC 61508 of the limit switch

SE - SIL conformity according to IEC 61508 of the current output

* positions which are not needed are omitted (no blank positions)

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

15

3

INSTALLATION

DK32 - DK34 - DK37

3.1 General notes on installation

INFORMATION!

Inspect the packaging carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.

INFORMATION!

Do a check of the packing list to make sure that you have all the elements given in the order.

INFORMATION!

Check on the device nameplates, that the device is supplied according to your order.

3.2 Storage

• Store the device in a dry, dust-free location.

• Avoid lasting direct exposure to the sun.

• Store the measuring device in the original packaging.

• The permissible storage temperatures are: -40...+80 ° C / -40...+176 ° F

3.3 Transport

If possible, transport the device to the installation point in the original packaging.

3.4 Installation conditions

CAUTION!

When installing the device in the piping, the following points must be observed:

• The variable area flowmeter must be installed vertically (measuring principle).

Flow direction from bottom to top.

Before connecting, blow or flush out the pipes leading to the device.

Piping for gas flow need to be dried before the device is installed.

Use connectors suitable for the particular device version.

Align the piping centrically with the connection bores on the measuring device so they are free of stresses.

If necessary, the piping has to be supported to avoid the vibrations transmitted to the measuring device.

Do not lay signal cables directly next to cables for the power supply.

When several devices are installed next to each other, a minimum distance between the devices is necessary (for details refer to chapter "Technical data").

The device must not be heated by radiated heat (e.g. exposure to the sun) to an electronics housing surface temperature above the maximum permissible ambient temperature. If it is necessary to prevent damage from heat sources, a heat protection (e.g. sun shade) has to be installed.

16 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

ELECTRICAL CONNECTIONS

4

4.1 Safety instructions

DANGER!

All work on the electrical connections may only be carried out with the power disconnected.

Take note of the voltage data on the nameplate!

DANGER!

Observe the national regulations for electrical installations!

DANGER!

For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.

WARNING!

Observe without fail the local occupational health and safety regulations.

Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists.

INFORMATION!

Look at the device nameplate to ensure that the device is delivered according to your order.

Check for the correct supply voltage printed on the nameplate.

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

17

4

ELECTRICAL CONNECTIONS

DK32 - DK34 - DK37

4.2 Electrical connection of limit switches

4.2.1 Electrical connection DK32 / DK34

The electrical connection of the limit switches is made for

• DK../../S - in connection plug

• DK../../L - via a cable assembly

The following procedures are to be performed (DK../../S):

• Loosen the screw 6 of the connector plug

• Pull out the plug.

• Remove the screw 6 completely from the plug.

• Insert a screwdriver in the marked opening 5 (lift) and remove the terminal block.

• Lead the connection cable through the cable gland.

• Insert the cable (max. 1.5 mm

2

) and tighten it.

Figure 4-1: Electrical connection of NAMUR limit switches

5 Lift slot

6 Fastening screw of connection box

Contact connection

1

2

3

4

Min minus

Min plus

Max minus

Max plus

Table 4-1: Contact connections

Stranded wire colour for cable assembly white yellow green brown

18 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

ELECTRICAL CONNECTIONS

4

Figure 4-2: Connection of 3-wire reed contact

1

2

3

Silicone-insulated wire - yellow/green / FEP-insulated wire - red

Silicone-insulated wire - brown / FEP-insulated wire - brown

Silicone-insulated wire - blue / FEP-insulated wire - blue

Table 4-2: Stranded wire colour for devices with cable assembly

4.2.2 Electrical connection DK37 M8M

Figure 4-3: Electrical connection of NAMUR limit switches

Contact

1

2

Kmin +

Kmin -

Kmax +

Kmax -

Table 4-3: Contact connections

Contact connection

Min plus

Min minus

Max plus

Max minus

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

19

4

ELECTRICAL CONNECTIONS

DK32 - DK34 - DK37

4.2.3 Settings of NAMUR limit switches for DK32, DK34, DK37 M8M

Set the pointers to the desired limit values as a min. contact 1 or max. contact 2 along the scale.

Figure 4-4: Settings of NAMUR limit switches for DK32, DK34, DK37 M8M

If the pointer vane goes into the slot, an alarm is triggered. An alarm is also triggered in the event of a cable break or short circuit if a suitable amplifier is used.

4.2.4 Settings of Reed limit switch

20

Figure 4-5: Settings of Reed limit switch

• Loosen the nut 1 .

• Set the reed cartridge to the desired value.

• Fix the nut 1 .

INFORMATION!

The reed contact is actuated directly by the float magnet. The desired switching point can only be determined in measuring mode. A reference to the scale and/or pointer cannot be established.

www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

ELECTRICAL CONNECTIONS

4

4.3 Electrical signal output ESK3x for DK32, DK34

Figure 4-6: Electrical connection of ESK3x

1 Lock screw in case of Ex d/XP device version

2 Terminal block

3 Power supply 12...32 VDC

4 Measuring signal 4...20 mA

5 External load, HART ® communication

The following procedures are to be performed:

• Loosen the Ex d/XP lock screw 1 on the locking mechanism using an Allen key (WS3).

• Unscrew the cover.

• Connect the connecting cables to the device terminals in the terminal block 2 by ensuring the correct polarity.

• Screw on the cover.

• Tighten the Ex d/XP lock screw on the locking mechanism using an Allen key (WS3).

DANGER!

For devices intended for use in hazardous areas with dust-resistant housing and/or secondary pressure containment, the special conditions for sealing the electronic compartment and for the cable entry are available in the supplementary instructions.

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

21

4

ELECTRICAL CONNECTIONS

DK32 - DK34 - DK37

4.3.1 Power supply

INFORMATION!

The supply voltage has to be between 12 VDC and 32 VDC. This is based on the total resistance of the measuring loop. To calculate this, the resistance of each component in the measuring loop

(not including the device) must be added up.

The required supply voltage can be calculated using the following formula:

U ext

= R

L

* 22 mA + 12 V with

U ext

= minimum supply voltage

R

L

= total measuring loop resistance

INFORMATION!

The power supply has to be able to supply a minimum of 22 mA.

4.3.2 Load for HART communication

INFORMATION!

For HART ® communication a load of at least 250 Ω is required.

The maximum load resistance is calculated as follows:

R

L

= (U ext

- 12 V) / 22 mA

DANGER!

Use a twisted two-core cable to prevent electrical interference from impeding the DC output signal.

In some cases a shielded cable may be necessary. The cable shield may only be grounded

(common ground connection) at one point (on the power supply unit).

4.3.3 Configuration

The ESK3x electrical signal output can be configured via HART ® communication. DDs (Device

Description) and DTMs (Device Type Manager) are available for the configuration.

They can be downloaded free of charge from our website.

The current flow rate can be transmitted using the integrated HART ® communication.

A flow counter can be configured. Two limit values can be monitored. The limit values are assigned either to flow values or to the counter overflow.

22 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

4.4 Electrical signal output for DK37/M8E

ELECTRICAL CONNECTIONS

4

Figure 4-7: Electrical connection M8E

1 Terminal connection

2 Power supply 14.8...30 VDC

3 Measuring signal 4...20 mA

4 External load, HART ® communication

Wiring must be planned with great care when it comes to connecting other devices such as evaluation units or process control. Internal connections in these devices (e.g. GND with PE, mass loops) may lead to non-permitted voltage potentials which could negatively affect the function of the device itself or that of a device connected to it. In such cases a protected extralow voltage (PELV) is recommended.

Figure 4-8: Electrical connection with galvanic isolation

1 Terminal connection

2 Signal converter isolator with galvanic isolation

3 Power supply (refer to isolator information)

4 Measuring signal 4...20 mA

5 External load, HART ® communication

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

23

4

ELECTRICAL CONNECTIONS

DK32 - DK34 - DK37

4.4.1 Power supply

INFORMATION!

The supply voltage has to be between 14.8 VDC and 30 VDC. This is based on the total resistance of the measuring loop. To calculate this, the resistance of each component in the measuring loop

(not including the device) must be added up.

The required supply voltage can be calculated using the following formula:

U ext

= R

L

* 22 mA + 14.8 V with

U ext

= minimum supply voltage

R

L

= total measuring loop resistance

INFORMATION!

The power supply has to be able to supply a minimum of 22 mA.

4.4.2 Load for HART communication

INFORMATION!

For HART ® communication a load of at least 250 Ω is required.

The maximum load resistance is calculated as follows:

R

L

= (U ext

- 14.8 V) / 22 mA

DANGER!

Use a twisted two-core cable to prevent electrical interference from impeding the DC output signal.

In some cases a shielded cable may be necessary. The cable shield may only be grounded

(common ground connection) at one point (on the power supply unit).

4.4.3 Configuration

The M8E electronic indicator can be configured via HART ® communication. DDs (Device

Description) and DTMs (Device Type Manager) are available for the configuration.

They can be downloaded free of charge from our website.

The current flow rate can be transmitted using the integrated HART ® communication.

A flow counter can be configured. Two limit values can be monitored. The limit values are assigned either to flow values or to the counter overflow. The limit values are not depicted on the display.

24 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

ELECTRICAL CONNECTIONS

4

4.5 Grounding connections

The measuring unit is grounded on the top fitting 1 (M4 threaded holes).

Alternatively, the DK32 and DK34 with ESK3x transmitter can be grounded at the transmitter housing.

Figure 4-9: Grounding connections

1 Grounding connection at measuring unit

DANGER!

The grounding wire may not transfer any interference voltage.

Do not use this grounding cable to ground any other electrical devices.

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

25

4

ELECTRICAL CONNECTIONS

DK32 - DK34 - DK37

4.6 Ingress protection

The variable area flowmeters DK32 / DK34 / DK37 have been tested for IP protection in accordance with EN 60529 and NEMA 250.

The following protection categories have been achieved:

Device variant

DK32, DK34 mechanical

DK32, DK34 with limit switches K1/K2

DK32, DK34 with limit switch R1

DK32, DK34 with transmitter ESK3x

DK37 with PPS indicator

DK37 with stainless steel indicator

Table 4-4: Ingress protection

Ingress protection according to

EN 60529

IP66/IP68

IP65

IP65

IP66/IP68

IP66

IP66/IP67

NEMA 250

NEMA 4X

NEMA 4X

-

NEMA 4X

-

NEMA 4X

DANGER!

After all servicing and maintenance work on the measuring device, the specified ingress protection category must be ensured again.

Therefore it is essential to observe the following points:

• Use only original gaskets. They must be clean and free of any damage. Defective gaskets must be replaced.

• The electrical cables used must be undamaged and must comply with regulations.

• The cables must be laid with a loop 3 upstream of the measuring device to prevent water from getting into the housing.

• The cable feedthroughs 2 must be tightened.

• Close the unused cable feedthroughs with blanking plugs 1 .

• Do not remove the required cable bushing from the cable feedthrough.

26

Figure 4-10: Cable feedthrough

1 Use blanking plugs if no cable is routed through

2 Tighten the protection cap

3 Lay the cable in a loop www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

START-UP

5

5.1 Start-up

CAUTION!

When starting up the device, the following points must be observed:

• Compare the actual operating pressure and the product temperature of the system with the specifications on the nameplate (PS and TS). These specifications may not be exceeded.

Make sure materials are compatible.

Slowly open the shut-off valve.

When measuring liquids ensure that the pipelines are carefully evacuated.

When measuring gases, increase pressure slowly.

Avoid float impact (e.g. caused by solenoid valves), as this is likely to damage the measuring unit or float.

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

27

6

SERVICE

DK32 - DK34 - DK37

6.1 Maintenance

Within the scope of routine maintenance of the system and pipelines, the flowmeter should also be inspected for signs of fouling, corrosion, mechanical wear and leaks, as well as damage to the measuring tube and the indicator.

We advise that inspections are carried out at least once per year.

The device must be removed from the pipeline before cleaning.

CAUTION!

Pressurised pipes have to be depressurised before removing the device.

In the case of devices used for measuring aggressive or hazardous media, appropriate safety precautions must be taken with regard to residual liquids in the measuring unit.

New gaskets have to be used when re-installing the device in the piping.

Figure 6-1: Valve lock

CAUTION!

The packing seal for the valve may have to be adjusted during the service life. The union nut 1 must be tightened for this. It may be necessary to push the locking pin 2 back against its internal spring.

Comply with a maximum torque of 4 Nm.

CAUTION!

Valves that have not been used for a longer period of time may exhibit a higher actuating torque.

28 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

SERVICE

6

6.2 Spare parts availability

The manufacturer adheres to the basic principle that functionally adequate spare parts for each device or each important accessory part will be kept available for a period of 3 years after delivery of the last production run for the device.

This regulation only applies to spare parts which are subject to wear and tear under normal operating conditions.

6.3 Availability of services

The manufacturer offers a range of services to support the customer after expiration of the warranty. These include repair, maintenance, technical support and training.

INFORMATION!

For more precise information, please contact your local sales office.

6.4 Returning the device to the manufacturer

6.4.1 General information

This device has been carefully manufactured and tested. If installed and operated in accordance with these operating instructions, it will rarely present any problems.

WARNING!

Should you nevertheless need to return a device for inspection or repair, please pay strict attention to the following points:

• Due to statutory regulations on environmental protection and safeguarding the health and safety of the personnel, the manufacturer may only handle, test and repair returned devices that have been in contact with products without risk to personnel and environment.

• This means that the manufacturer can only service this device if it is accompanied by the following certificate (see next section) confirming that the device is safe to handle.

WARNING!

If the device has been operated with toxic, caustic, radioactive, flammable or water-endangering products, you are kindly requested:

• to check and ensure, if necessary by rinsing or neutralising, that all cavities are free from such dangerous substances,

• to enclose a certificate with the device confirming that it is safe to handle and stating the product used.

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

29

6

SERVICE

DK32 - DK34 - DK37

6.4.2 Form (for copying) to accompany a returned device

CAUTION!

To avoid any risk for our service personnel, this form has to be accessible from outside of the packaging with the returned device.

Company:

Department:

Address:

Name:

Telephone number:

Fax number:

Manufacturer order number or serial number:

Email address:

The device has been operated with the following medium:

This medium is: radioactive water-hazardous toxic caustic flammable

We checked that all cavities in the device are free from such substances.

We have flushed out and neutralized all cavities in the device.

We hereby confirm that there is no risk to persons or the environment caused by any residual media contained in this device when it is returned.

Signature: Date:

Stamp:

6.5 Disposal

LEGAL NOTICE!

Disposal must be carried out in accordance with legislation applicable in your country.

According to the directive 2012/19/EU, the monitoring and control instruments marked with the

WEEE symbol and reaching their end-of-life must not be disposed of with other waste

The user must dispose of the WEEE to a designated collection point for the recycling of WEEE or send them back to our local organisation or authorised representative.

30 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

SERVICE

6

6.6 Disassembly and recycling

This section describes how to handle and disassemble the device in order to dispose of it following end of use. The information provided enables the end user to identify and separate the main components of the device for the purpose of recycling.

INFORMATION!

• Wear personal protective equipment

• Make sure to use a stable workstation / bench to disassemble

6.6.1 Description of the device components

Figure 6-2: Description of the device components

1 Measuring unit

2 Indicator housing

The variable area flowmeter consists of a mechanical, metallic measuring unit 1 and an indicator housing attached to it 2 , which may contain electronic components.

The metallic measuring unit can be sent to metal recycling after the indicator housing has been removed. Depending on the version, the measuring unit may weigh between 0.5...5 kg /

1.1...11 lb.

CAUTION!

Check and ensure that the measuring units, including all cavities, are free of hazardous substances by rinsing or neutralising if required. This should be done especially in the case of devices that have been operated with toxic, corrosive, radioactive, flammable or watercontaminating products to protect both the environment and the health and safety of staff.

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

31

6

SERVICE

DK32 - DK34 - DK37

6.6.2 Indicator versions

The following discusses the indicator housing and optional electronics in more detail.

Indicator DK37/M8x and DK37/M8x/R

The indicator housing consists of the following components:

Approx. information Weight / size

DK37/M8x/R

Stainless steel housing

Glass panel

PPS (housing including cover)

Stainless steel (housing including cover)

Other metal parts as scale, pointer system, screws

Insert profile made of PA

45 cm

-

Table 6-1: Indicator DK37/M8x and DK37/M8x/R

² / 6.98 in

1550 g / 3.42 lb

50 g / 0.1 lb

40 g / 0.09 lb

²

DK37/M8x

Plastic housing

45 cm ² / 6.98 in ²

300 g / 0.66 lb

-

50 g / 0.1 lb

-

The indicator housing can be equipped either with an electronic transmitter DK37/M8E or with one or two limit switches DK37/M8/K1 or DK37/M8/K2. They become visible after opening the cover. These can be removed from the housing in order to send them to electronics recycling/disposal.

The electronic components consist of the following main components:

Approx. information Weight / size

DK37/ … /M8E

PCB-CPU potted

PCB-I/O not potted

PCB connection board

Potting (silicone)

Batteries

Electrolytic capacitors

LC-display

67 cm

45 cm

17 cm ² / 2.64 in ²

95 g / 0.21 lb

-

-

24 cm

²

²

²

/ 10.39 in

/ 6.98 in

/ 3.72 in

Table 6-2: Main components of the electronic components

²

²

²

DK37/ … /M8M/K.

< 1 cm

²

/ < 0.16 in

²

-

-

-

5 / 10 cm ² / 0.78 / 1.55 in ²

5 g / 0.01 lb

-

INFORMATION!

The PCBs are partially potted with silicone. It is thus not easy to separate the components by hand.

32 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

SERVICE

6

Indicator DK32, DK34

The indicator housing consists of the following components:

Approx. information Weight / size

DK3./K.

Glass panel

Aluminium housing

Other metal parts as scale, pointer system, screws

Table 6-3: Indicator DK3./K. and DK3./ESK

9 cm ² / 1.40 in ²

170 g / 0.37 lb

50 g / 0.1 lb

DK3./ESK

9 cm ² / 1.40 in ²

460 g / 1.01 lb

50 g / 0.1 lb

The indicator housing can be equipped either with an electronic transmitter DK3./ESK or with one or two limit switches DK3./K. These can be removed from the housing in order to send them to electronics recycling/disposal.

The electronic components consist of the following main components:

Approx. information Weight / size

DK3./K.

PCB

Potting (silicone)

Batteries

Electrolytic capacitors

LC-display

1 cm

5 g / 0.01 lb

-

-

-

² / 0.16 in ²

Table 6-4: Main components of the electronic components

DK3./ESK

-

-

25 cm ² / 3.88 in ²

28 g / 0.06 lb

-

INFORMATION!

The PCBs are potted with silicone. It is thus not easy to separate the components by hand.

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

33

7

TECHNICAL DATA

DK32 - DK34 - DK37

7.1 Functional principle

The flowmeter operates in accordance with the float measuring principle.

The measuring unit consists of a metal cone in which a float can move freely up and down.

The medium flows through the flowmeter from bottom to top.

The float adjusts itself so that the buoyancy force A acting on it, the form resistance W and its weight G are in equilibrium: G = A + W

Figure 7-1: Functional principle

For DK32, DK34 and DK37/M8M 1 the flow-dependent height of the float in the measuring unit is transmitted by means of a magnetic coupling on a scale.

With DK37/M8E, DK32/ESK and DK34/ESK 2 the flow-dependent height position of the float in the measuring unit is electronically detected with the magnetic field sensors S1 and S2.

34 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

TECHNICAL DATA

7

7.2 Technical data

INFORMATION!

• The following data is provided for general applications. If you require data that is more relevant to your specific application, please contact us or your local sales office.

• Additional information (certificates, special tools, software,...) and complete product documentation can be downloaded free of charge from the website (Downloadcenter).

Measuring system

Application range

Function / Measuring principle

Measured value

Primary measured value

Secondary measured value

Measuring accuracy

Maximum measuring error DK32 / DK34

Maximum measuring error DK37

Operating conditions

Max. operating temperature TS

Flow measurement of liquids or gases

Float measuring principle

Float position

Operating volume flow, nominal volume flow or mass flow

3.0% of full scale range

4.0% of measured value according to VDI/VDE 3513-2 (q

G

= 50%)

2.0% of full scale range

2.5% of measured value according to VDI/VDE 3513-2 (q

G

= 50%)

Max. operating pressure PS

Max. test pressure PT

Depending on the version (refer to nameplate)

Standard version: up to +150 ° C / +221 ° F

HT version: up to +200 ° C / +392 ° F

For detailed information on product and ambient temperatures refer to Temperatures for all device versions on page 37.

Depending on the version (refer to nameplate)

Standard version: 130 barg / 1885 psig

Higher pressures on request

Depending on the version (refer to nameplate)

Standard version: 202 barg / 2930 psig

Higher pressures on request

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

35

7

TECHNICAL DATA

DK32 - DK34 - DK37

DK32, DK34 mechanical

DK32, DK34 with limit switches K1/K2

DK32, DK34 with limit switch R1

DK32, DK34 with transmitter ESK3x

DK37 with PPS indicator

DK37 with stainless steel indicator

Installation conditions

Inlet and outlet sections

Protect devices from vibration and highfrequency oscillation

Materials

Top fitting, bottom fitting, cone

Lock screw

Float, standard

Dosing unit

Valve stem

Gasket of lock screw

Gasket of dosing unit

Indicator housing of DK32, DK34

Indicator housing of DK37/M8_/

Indicator housing of DK37/M8_/R/

Special materials

Options

Process connections

Standard

Option

Table 7-1: Technical data

IP66/IP68 and NEMA 4X

IP65 and NEMA 4X

IP65

IP66/IP68 and NEMA 4X

IP66

IP66/IP67 and NEMA 4X

None

Use in accordance with IEC 61298-3 in control room or field with medium vibration level.

Stainless steel 1.4404 / 316 L

Stainless steel 1.4404 / 316 L

Stainless steel 1.4404 / 316 L or titanium

Stainless steel 1.4571 / 316 Ti

Stainless steel 1.4404 / 316 L

PTFE

FPM/PTFE, PTFE, FFKM

Other gasket materials on request

Die-cast aluminium with powder coating

PPS

Stainless steel 1.4408/CF8M e.g. Titanium Grade 2, Hastelloy ® C276/2.4819, Monel ® /2.4360,

Inconel ® /2.4856,...

Surface passivation of all metallic, wetted parts e.g. SilcoNert ® 2000 or Dursan ®

1/4 NPT female

G1/4, Ermeto, Serto, Gyrolok, Swagelok, flanges

Other connections on request

36 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

TECHNICAL DATA

7

7.2.1 Temperatures for all device versions

DANGER!

For devices to be used in hazardous areas, special temperature ranges apply. These can be found in the Ex supplementary instructions.

Device variant

Max. medium temperature

Permissible ambient temperatures

[ ° C]

150 100

DK32, DK34, DK37/M8M without electrical built-ins

DK32, DK34, DK37/M8M with limit switch I7S2002-N or SJ2-SN

DK32, DK34, DK37/M8M with limit switch SC2-N0

DK32, DK34 with transmitter ESK3x

DK37/M8E

-40

-40

-25

+70

+40

+40

-40 … +40

-40 … +40

-40

-40

-40

-40

-40

+90

+90

+90

+70

+70

Table 7-2: Permissible ambient temperatures for electronics variants in ° C and ° F

[ ° F]

302

-40...+158

-40...+104

-25...+104

-40...+104

-40...+104

212

-40...+194

-40...+194

-40...+194

-40...+158

-40...+158

Device variant Permissible medium temperatures

[ ° C]

DK34, DK32, DK37 without valve

DK32, DK37 with valve FPM/PTFE

DK32, DK37 with valve FFKM

DK32, DK37 with valve PTFE

-80

-25

-20

-40

+150

+150

+150

+150

Table 7-3: Permissible medium temperatures for gasket variants in ° C and ° F

[ ° F]

-112...+302

-13...+302

-4...+302

-40...+302

Device variant Permissible medium and ambient temperatures

[ ° C] [ ° F]

Max. ambient temperature 150 180 302 356

DK32 HT with FPM/PTFE, DK34 HT

(optional Reed contact)

DK32 HT with FFKM

(optional Reed contact)

-25

-20

+150

+200

-

-20

+180

-13...+302

-4...+392

-

-4...+356

Table 7-4: Permissible medium and ambient temperatures for high-temperature version DK32 HT/DK34 HT in ° C and ° F

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

37

7

TECHNICAL DATA

DK32 - DK34 - DK37

7.2.2 Electrical connection data for indicator DK32, DK34, DK37/M8M with limit switch

DK32 - 34 /K./S

DK37/M8M

DK32 - 34 /K./S

DK37/M8M

DK32 - 34/K./L

Cable gland: M16 x 1.5

Terminal connection: 1.5 mm

2

Clamping range: 4.5...10 mm / 0.18...0.4"

Clamping range: 3...7 mm / 0.12...0.28"

Ø Connecting cable: 6.3 mm / 0.25"; 0.75 mm ²

Cable length: approx. 1.7 m / 5.6 ft (other lengths on request)

Limit switch

SIL2-compliant according to IEC 61508

Connection type NAMUR

Switching element function

Nominal voltage U

0

Pointer vane not detected

Pointer vane detected

DK32 - DK34 with Reed contact no

2-wire

Normally closed

8 VDC yes

2-wire

Normally closed

8 VDC yes

2-wire

Normally closed

8 VDC

≥ 3 mA

≤ 1 mA

3 mA

1 mA

3 mA

1 mA

Type: SPDT (3-wire, normally closed and normally open)

Switching reproducibility: < 5% of full scale range

Switching capacity: 12 VA 1

Max. switching capacity: 30 VDC 1

Max. switching current: 0.5 A 1

Table 7-5: Electrical connection data for indicator with limit switch

1 Reduced values for Ex

38 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

TECHNICAL DATA

7

7.2.3 Electrical connection data for transmitter ESK3x for DK32, DK34

Cable entry

Clamping range of cable gland

Terminal connection

Measuring signal

Power supply

Power supply influence

Dependence on load impedance

Temperature influence

Max. external resistance / load

Min. load for HART ®

NAMUR type test acc. to NE 95

M16 x 1.5 or 1/2" NPT

3...7 mm / 0.12...0.28" (PA gland) or 5...10 mm / 0.2...0.39" (stainless steel gland)

1.5 mm

2

4...20 mA = 0 … 100% flow value in 2-wire technology

12...32 VDC (non-Ex)

< 0.1%

< 0.1%

<0.3%/10 K

500 Ω (24 VDC), 800 Ω (32 VDC)

250 Ω

Fulfills NE 21, NE 43, NE 107

ESK3x - electrical characteristics

Process variable flow rate

Upper range limit

Device error identification

Maximum

Multidrop operation

Min. U ext.

+103.125

< -2.5 or >106.25

112.5

-

12 VDC

ESK3x HART Device

Manufacturer name (code)

Model name

KROHNE Messtechnik (0x0045)

ESK3x (0x45BA)

Revision 7.6

HART ® protocol

Device revision

Equipment category

1

Transmitter

Table 7-6: Electrical connection data for transmitter ESK3x

20.5

< 3.6 or > 21

22.0

4.0

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

39

7

TECHNICAL DATA

DK32 - DK34 - DK37

7.2.4 Electrical connection data for indicator DK37/M8E with electrical signal output

Cable gland

Clamping range

Terminal connection

Measuring signal

Power supply

Min. power supply for HART ®

Power supply influence

Dependence on load impedance

Temperature influence

Max. external resistance / load

Min. load for HART ®

Software firmware version

Identification number

NAMUR type test acc. to NE 95

M16 x 1.5

3...7 mm / 0.12...0.28"

2.5 mm

2

4...20 mA = 0 … 100% flow value in 2-wire technology

14.8...30 VDC

20.5 VDC

<0.1%

<0.1%

<10

µ

A/K

640 Ω (30 VDC)

250 Ω

01.15

3204090400

Fulfills NE 21, NE 43

M8E HART

®

configuration

Manufacturer name (code)

Model name

HART ® protocol

Device revision

Physical Layer

Equipment category

KROHNE Messtechnik (69)

M8E (230)

Revision 5.1

1

FSK

Transmitter

M8E process variable

Process variable flow rate

Upper range limit

Device error identification

Maximum

Multidrop operation

Min. U ext.

+102.5 ( ± 1%)

>106.25

112.5

-

14.8 VDC

Table 7-7: Electrical connection data for indicator with electrical signal output

20.24...20.56

≥ 21.00

22

4.5

40 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

TECHNICAL DATA

7

7.2.5 Approvals

Standard Device version Marking

ATEX / IECEx

QPS (US/C)

Mechanical, without electrical built-ins II2G Ex h IIC T6...T3 Gb

II2D Ex h IIIC T150

°

C Db

NAMUR limit switch K1/K2 II2G Ex ia IIC T6 Gb

II2D Ex ia IIIC T75 ° C Db

II3G Ex nA/ec IIC T6 Gc

Reed limit switch R1

4...20 mA / HART signal output ESK3x

NAMUR limit switch K1/K2

II2G Ex ia IIC T6 Gb

II2G Ex d IIC T6 Gb

II2G Ex ia IIC T6 Gb

II2D Ex ia IIIC T75 ° C Db

II3G Ex nA/ec IIC T6 Gc

II2D Ex tb IIIC 75 ° Db

IS Class I,II,III Div 1 / Zone 1/21 AEx ia/Ex ia

NI Class I Div 2 / Zone 2 AEx nA/Ex nA

Reed limit switch R1

4...20 mA / HART signal output ESK3x

IS Class I Div 1 / Zone 1 AEx ia/Ex ia

IS Class I,II,III Div 1 / Zone 1/21 AEx ia/Ex ia

NI Class I Div 2 / Zone 2 AEx nA/Ex nA

XP Class I Div 1 / Zone 1 AEx db/Ex db

DIP Class II, III Div 1 / Zone 21 AEx tb/Ex tb

Other approvals as NEPSI, INMETRO, KCS, CCOE, EAC on request.

Table 7-8: Approvals for DK32, DK34

Standard Device version Marking

ATEX / IECEx

QPS (US/C)

Mechanical, without electrical built-ins II2G Ex h IIC T6...T3 Gb

II2D Ex h IIIC T150 ° C Db

Electrical II2G Ex ia IIC T6 Gb

II2D Ex ia IIIC T75 ° C Db

Electrical IS/I/1/ABCD; T6

Other approvals as NEPSI, INMETRO, KCS, CCOE, EAC on request.

Table 7-9: Approvals for DK37

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

41

7

TECHNICAL DATA

DK32 - DK34 - DK37

7.3 Dimensions and weight

7.3.1 DK32, DK34

DK32 with top/bottom valve and rear process connection DK34 without valve and vertical process connection

1

Version with limit switch with connector

DK32 with bottom valve /

DK34 without valve

DK32 with top valve

2 a

[mm] ["]

1

2

DK32

DK34

42

42

Table 7-10: Dimensions and weight

1.66

1.66

Dimensions

[mm]

118

110 b

["]

4.65

4.33

[mm]

90

c

["]

3.55

-

[mm]

100

75 d

["]

3.94

3.07

g

700

600

Weight approx.

Version with limit switch with cable end

DK32 with bottom valve /

DK34 without valve

DK32 with top valve lb

1.54

1.32

3 a

[mm] ["]

3 DK32

DK34

46

Table 7-11: Dimensions and weight

1.81

b

Dimensions

[mm] approx.

90

["] approx.

3.55

[mm]

1500 c

Weight for DK32 with flow regulator: 2500 g / 5.51 lb

["]

59 d

[mm] approx.

50

["] approx.

1.97

g

700

600

Weight approx.

lb

1.54

1.32

42 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

Version with flange adapter

DK32 with bottom valve DK32 with top valve

TECHNICAL DATA

7

DK34 without valve

4

Dimensions

4 DK32

[mm]

-

DK34 -

Table 7-12: Dimensions and weight a

["]

-

-

[mm]

250

250 b

["]

10.2

10.2

[mm]

90

c

["]

3.55

-

[mm] approx.

195

d

["] approx.

7.68

-

Extra weight for flange adapter dependent on pressure rating of flange: approx. 1.5...9 kg /

3.3...19.8 lb

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

43

7

TECHNICAL DATA

7.3.2 DK32, DK34 with transmitter housing ESK3x

DK32 with ESK transmitter housing with top/bottom valve

DK32 - DK34 - DK37

1 a b

[mm]

1 DK32

Table 7-13: Dimensions

100

["]

3.94

[mm]

118

Total length of flange connection adapter:

DK32/ESK with top or bottom valve: 250 mm / 9.8"

["]

4.65

DK32 with transmitter ESK3x

DK34 with transmitter ESK3x

Table 7-14: Weight approx. g

1100

1000

[mm]

90 c

["]

3.55

approx. lb

2.43

2.21

[mm]

140 d

["]

5.51

44 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

TECHNICAL DATA

7

7.3.3 DK37/M8M

With bottom valve and rear process connection

With top valve and rear process connection

Without valve and vertical process connection

1 a b

1

2

With valve

Without valve

[mm]

56

56

["]

2.21

2.21

[mm]

153

145

Table 7-15: Dimensions

Total length of flange connection adapter:

DK37/M8M with top/bottom valve: 250 mm / 9.8"

["]

6.03

5.71

DK37/M8M

DK37/M8M with flow regulator

DK37/M8M/R

DK37/M8M/R with flow regulator

Table 7-16: Weight approx. g

800

2600

2100

3900

2

[mm]

125

145 c

["]

4.92

5.71

[mm] approx.

144 approx.

144 d

["] approx.

5.67

approx.

5.52

approx. lb

1.76

5.73

4.63

8.60

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

45

7

TECHNICAL DATA

DK32 - DK34 - DK37

7.3.4 DK37/M8E

With bottom valve and rear process connection

With top valve and rear process connection

Without valve and vertical process connection

1

DK37/M8E

DK37/M8E with flow regulator

DK37/M8E/R

DK37/M8E/R with flow regulator

Table 7-18: Weight

2 a b

1

2

3

With bottom valve

With top valve

Without valve

[mm]

56

56

56

["]

2.21

2.21

2.21

[mm]

153

183

145

Table 7-17: Dimensions

Total length of flange connection adapter:

DK37/M8E with bottom valve: 250 mm / 9.8"

DK37/M8E with top valve: 300 mm / 11.8"

["]

6.03

7.21

5.71

approx. g

1000

2800

2300

4100

3

[mm]

125

155

145 c

["]

4.92

6.11

5.71

[mm] approx.

144 approx.

144 approx.

121 d

["] approx.

5.67

approx.

5.67

approx.

4.77

approx. lb

2.21

6.18

5.07

9.04

46 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

TECHNICAL DATA

7

7.3.5 Minimum distances when installing several measuring devices

When installing several measuring devices side by side, a minimum distance between them has to be observed.

Figure 7-2: Minimum distances when installing several measuring devices

Device [mm]

1

2

3

3

DK32, DK34

DK32/ESK, DK34/ESK

DK37/M8M

DK37/M8E

Table 7-19: Minimum distances in mm and inch

60

140

100

120

["]

2.36

5.51

3.94

4.73

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

47

7

TECHNICAL DATA

DK32 - DK34 - DK37

7.4 Measuring ranges

7.4.1 Float

Measuring span:

Flow values:

10 : 1

Values = 100%; Water: +20 ° C / +68 ° F; Air: +20 ° C / +68 ° F, 1.013 bara / 14.7 psia

Cones Float

K 040

K 080

K 080

K 125

K 125

K 200

K 300

K 005

K 005

K 005

K 010

K 010

K 015

K 015

K 040

K 340

K 340

K 340 2

K 340 2

Table 7-20: Measuring ranges for floats

N3 titanium 1

N1 titanium 1

N1 stainless steel

N1 titanium 1

N1 stainless steel

N1 titanium 1

N1 stainless steel

N2 titanium 1

N2 stainless steel

N2 titanium 1

N2 stainless steel

N2 titanium 1

N2 stainless steel

N2 stainless steel

N2 stainless steel

N2 stainless steel

N2 Triamet

Special stainless steel

Special Triamet

Flow, water

[l/h]

40

60

80

10

16

25

25

100

150

200

250

5

7

3

3

-

-

-

1.5

[GPH]

11

16

21

2.6

4.2

6.6

6.6

26

40

53

66

0.8

0.8

1.3

1.8

-

-

-

0.4

Flow, air

[Nl/h]

400

550

800

-

1250

2000

2500

3400

4800

6000

8000

100

100

150

250

16

25

50

70

[SCFH]

50

75

100

15

20

30

-

125

180

225

300

6

10

4

4

0.6

1

2

2.6

Pressure loss

[mbar]

17

42

85

117

27

32

55

45

166

210

462

722

31

31

38

17

19

17

[psi]

0.25

0.39

0.47

0.8

0.65

0.61

1.23

1.7

0.45

0.45

0.55

0.25

0.28

0.25

2.4

3.01

6.7

10.5

1 Titanium components are not suitable for use in oxygen applications (media with an oxygen content that is significantly higher than the oxygen content in the earth's atmosphere)!

2 DK37 only

INFORMATION!

The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI/VDE directive 3513.

Flow measurements for gases are attributed to:

- Nl/h or Nm

3

/h: Volume flow at standard (norm.) conditions 0 ° C / +32 ° F, 1.013 bara / 14.7 psia

(DIN 1343)

- SCFM or SCFH: Volume flow at standard (std.) conditions +15 ° C / +59 ° F, 1.013 bara / 14.7 psia

(ISO 13443)

48 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

TECHNICAL DATA

7

7.4.2 Valves (only DK32 and DK37)

Measuring span:

Flow values:

10 : 1

Values = 100%; Water: +20 ° C / +68 ° F; Air: +20 ° C / +68 ° F, 1.013 bara / 14.7 psia

Cones

Valve stem

Ø [mm] Ø ["] [l/h]

K 005 - K 010

K 015 - K 040 - K 080

K 125 - K 340

1

2.5

4.5

0.039

0.98

0.177

5

50

160

Table 7-21: Measuring ranges for valves (only DK32 and DK37)

Water

[GPH]

1.32

13.2

42.3

Max. flow Qv

[Nl/h]

100

1000

4300

Air

[SCFH]

3.72

37.2

160

Valve characteristic value

Kv Cv

[GPM]

[m

3

/h]

0.018

0.15

0.48

0.021

0.175

0.552

Table 7-22: Valve characteristics

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

1 Flow, air

2 Flow, water

3 Spindle revolution n

49

7

TECHNICAL DATA

DK32 - DK34 - DK37

7.5 Flow regulators for variable pressure (only DK32 and DK37)

Flow regulators are used (only DK32 and DK37) to guarantee constant flow during fluctuating inlet and outlet pressure. Minimum pressures are required to operate the regulators (refer to regulator characteristics).

INFORMATION!

Flow regulators are not equivalent to pressure regulators / pressure reducing valves!

1

Flow regulator for variable inlet pressure, type RE, NRE

The regulators keep the flow rate constant in the case of a variable inlet pressure and a constant outlet pressure.

RE-1000 Current flow rate:

Outlet pressure p2 constant:

Table 7-23: Example for flow regulator for variable inlet pressure

1000 l/h air

1.013 bara / 14.7 psia

The flow rate in the device remains constant with a fluctuating inlet pressure greater than 0.5 bar / 7.25 psi.

2

Flow regulator for variable outlet pressure, type RA, NRA

The regulators keep the flow rate constant in the case of a constant inlet pressure and a variable outlet pressure.

There must be a pressure differential between the inlet and the outlet pressure for the flow regulators to function. The inlet pressure p1 must always be greater than the outlet pressure p2.

RA-1000 Current flow rate: 800 l/h air

Inlet pressure p1 constant: 6 bar / 87 psi

Table 7-24: Example for flow regulator for variable outlet pressure

The flow rate in the device remains constant with a fluctuating outlet pressure greater than 0...5.5 bar / 0...79.8 psi.

1 Flow regulators for variable inlet pressure, type

RE and NRE

2 Flow regulators for variable outlet pressure, type RA and NRA

50

Table 7-25: Regulator characteristics www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

TECHNICAL DATA

7

7.5.1 Control ranges

Measuring span:

Flow values:

10 : 1

Values = 100%; Water: +20 ° C / +68 ° F; Air: +20 ° C / +68 ° F, 1.013 bara / 14.7 psia

Max. flow rate

Water Air

[l/h] [GPH] [Nl/h] [SCFH]

RE-1000

RE-4000

NRE-100

NRE-800

...40

...80

...100

...160

...2.5

-

-

...25

...11

...20

...25

...42

...0.6

-

-

...6.6

...1000

...2000

-

3000

...4000

...100

… 250

800

Table 7-26: Control ranges for flow regulator for variable inlet pressure 1

...37

...75

...110

...150

...3.7

...9.0

...30

-

Min. inlet pressure p1 [bar]

0.1

0.1

0.2

0.4

0.5

1

1.5

2 p1 [psi]

7.25

14.5

21.8

29

1.45

1.45

2.9

5.8

Max. flow rate Min. inlet pressure Min. pressure diff.

*

Water Air

[l/h] [GPH] [Nl/h] [SCFH] p1 [bar] p1 [psi] Δ p [bar] Δ p [psi]

RA-1000

RA-4000

NRA-800

...40

...100

-

...160

...1

-

-

...25

...11

...25

-

...42

...0.25

-

-

...6.6

...1000

...3000

...4000

...250

...500

-

2000

800

...37

...75

...110

...150

...9.0

...19

...30

-

Table 7-27: Control ranges for flow regulator for variable outlet pressure 2

0.5

1.5

1.5

2

0.1

0.2

0.4

0.4

7.25

21.8

21.8

29

1.45

2.9

5.8

5.8

0.4

1.2

1.2

1.5

0.05

0.1

0.2

0.4

5.8

17.4

17.4

21.8

0.73

1.45

2.9

5.8

* Differential pressure between inlet and outlet pressure

Flow measurements for gases are attributed to:

- Nl/h or Nm

3

/h: Volume flow at standard (norm.) conditions 0 ° C / +32 ° F, 1.013 bara / 14.7 psia

(DIN 1343)

- SCFM or SCFH: Volume flow at standard (std.) conditions +15 ° C / +59 ° F, 1.013 bara / 14.7 psia

(ISO 13443)

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

51

7

TECHNICAL DATA

DK32 - DK34 - DK37

7.5.2 Technical data for flow regulators

Connection, standard

Connection, option

Max. operating gauge pressure (at +20

°

C / +68

°

F)

Medium temperature

Material

Gasket

Diaphragm

O-ring

Table 7-28: Technical data

1/4 NPT

Serto, Ermeto 6 or 8, hose nozzle 6 mm or 8 mm, Dilo, Gyrolok,

Swagelok, G1/4

64 barg / 928 psig

+150 ° C / +302 ° F

Stainless steel 1.4404

PTFE

Carbon/graphite-filled PTFE

FPM or FFKM

INFORMATION!

Higher pressures and temperatures, other connections or materials on request.

52 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

TECHNICAL DATA

7

7.5.3 Dimensions with flow regulator

DK32 with flow regulator for variable inlet pressure DK37/M8M and M8E flow regulator for variable inlet pressure

DK32 with flow regulator for variable outlet pressure

DK37/M8E flow regulator for variable outlet pressure

[mm]

DK32

DK37

DK37/M8E 1

Table 7-29: Dimensions in mm and inch

230

230

230 approx. a

["]

9.1

9.1

9.1

1 With flow regulator for variable outlet pressure approx. b

[mm]

163

200

230

["]

6.4

7.9

9.1

[mm]

70

70

70 c

["]

2.8

2.8

2.8

[mm]

23

23

23 d

["]

0.91

0.91

0.91

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

53

8

NOTES

DK32 - DK34 - DK37

54 www.krohne.com

09/2020 - 4000287107 - MA DK32-34-37 R09 en

DK32 - DK34 - DK37

NOTES

8

09/2020 - 4000287107 - MA DK32-34-37 R09 en www.krohne.com

55

KROHNE – Process instrumentation and measurement solutions

• Flow

• Level

• Temperature

• Pressure

• Process Analysis

• Services

Head Office KROHNE Messtechnik GmbH

Ludwig-Krohne-Str. 5

47058 Duisburg (Germany)

Tel.: +49 203 301 0

Fax: +49 203 301 10389 [email protected]

The current list of all KROHNE contacts and addresses can be found at: www.krohne.com