Neptune APEX JR, APEX LITE Cloning Manual

Add to My manuals
17 Pages

advertisement

Neptune APEX JR, APEX LITE Cloning Manual | Manualzz

   

 

 

 

 

 

Table of Contents 

INTRODUCTION ......................................................................................................................................................... 3 

REQUIREMENTS ......................................................................................................................................................... 4 

ACCESSING THE CLASSIC DASHBOARD ...................................................................................................................... 7 

 

CLONING YOUR APEX .............................................................................................................................................. 10 

Create a configuration backup of the OLD APEX ................................................................................................. 10 

Prepare the NEW APEX ........................................................................................................................................ 11 

Restore the Configuration Backup on the NEW APEX ......................................................................................... 12 

Enable REPLACEMENT MODE in the NEW APEX ................................................................................................. 14 

Replace the OLD APEX with the NEW APEX ........................................................................................................ 16 

Disable REPLACEMENT MODE ............................................................................................................................. 16 

Additional Steps on the NEW APEX ..................................................................................................................... 16 

Link the NEW APEX with APEX Fusion ................................................................................................................. 17 

 

 

Page 2 

Version: 1.2 

 

 

INTRODUCTION 

 

 

On occasion, due to a failure or water damage it may become necessary to replace an APEX, APEX LITE, or APEX 

JR with another controller.      

In APEX‐family firmware version 4.40 and newer, a special mode has been added to facilitate cloning an APEX  configuration.  This is called REPLACEMENT MODE.  REPLACEMENT MODE, in conjunction with the Load/Save  features of the APEX, will allow you to quickly and easily swap one APEX base unit with another APEX base unit  or swap an APEX JR with another APEX JR, and preserve all existing programming, profiles, AquaBus module  addresses, and most other APEX settings.    

The cloning process and APEX REPLACEMENT MODE may only be used for direct replacement of an  identical  controller model.   It cannot be used as an upgrade or migration tool if changing from one model of APEX to  another, such as upgrading from an APEX JR to an APEX or upgrading from an APEX LITE to an APEX.   

 

Note:  The label on the face of the controller does not matter; an APEX or APEX JR with an old black label and an 

APEX or APEX JR with the newer white label both have the same hardware, so the label color is not a factor  when using this document. 

The procedures in this Guide are designed to cover a wide variety of different system configurations and home  or business network environments, and to accommodate Apex users of all experience and knowledge levels.   

Consequently, these procedures are very detailed, but please do not be intimidated or discouraged.   

Experienced APEX users can do this entire process in 15‐20 minutes; less‐experienced APEX users will likely  require a bit more time.   

NOTE:  Currently, it is not possible to migrate historical data stored in the Apex or in APEX Fusion from  one Apex to another.  This applies to:  

Logged probe measurement data 

Logged outlet activity data 

DŌS activity data 

 

Regardless of your experience level, it is  crucia l that these procedures be  followed exactly and that no steps be skipped or abbreviated.

 

 

 

Page 3 

Version: 1.2 

 

REQUIREMENTS 

There are three critical requirements which must be met before using this procedure and REPLACEMENT MODE: 

1.

You must already have, or be able to make a configuration backup file from the original APEX 

2.

You must have a functional APEX display module 

3.

The new APEX must be of the exact same model as the APEX being replaced 

There is one more requirement applicable only to APEX base units – this does not apply to APEX JR: 

4.

You must have a 12‐volt AC adapter for APEX 

 

Additionally, this process requires the use of a desktop or notebook computer running Windows or Mac OS.    

If you use an APEX but do not have a 12‐volt adapter: 

You may purchase one directly from Neptune Systems 

You may purchase one from your preferred reseller of Neptune Systems products 

You may acquire and use one from other sources if it meets these specifications:  o 12‐volt DC output  o 2.1mm plug  o Plug has positive (+) on the center pin  o DC output rating is 0.5 amps (500 milliamps), or greater 

Many home Internet routers, other home network gear, and some consumer electronics use similar AC  adapters; you might already have one on hand which may be used temporarily during the cloning  process. 

 

 

Use the flow chart on the next page to determine if you can use the cloning procedure. 

 

 

Page 4 

Version: 1.2 

 

Page 5 

START

Are the original

APEX and new APEX the 

SAME model?

NO

YES

STOP

You will need to manually  configure the new APEX.

Do you have a  functional display  module?

YES

Do you have a recent 

APEX configuration backup  file saved on your  computer?

YES

NO

NO

STOP

You cannot use the process

 

Version: 1.2 

 

Page 6 

Can you access  the 

original 

APEX’s Classic 

Dashboard?

NO

I don’t know how

YES

STOP

You will need to manually  configure the new APEX.

Go to the ACCESSING 

THE CLASSIC 

DASHBOARD section

Is this an APEX JR?

YES

NO

Do you  have a 12‐volt AC adapter  for your APEX?

YES

NO

Obtain a 12VDC AC  adapter, then continue

Go to the CLONING 

YOUR APEX section

Version: 1.2 

 

 

 

 

ACCESSING THE CLASSIC DASHBOARD 

DO NOT USE MOZILLA FIREFOX OR MICROSOFT EDGE BROWSERS DURING THE CLONING 

PROCESS 

Mozilla Firefox is fine to use for accessing an APEX’s Classic Dashboard in everyday use, but  during the APEX cloning process, the most recent version of Firefox will not work.  Microsoft 

Edge (included in Windows 10) also has known issues.  If you use a Windows computer, use 

Google Chrome or Microsoft Internet Explorer.  If you use an Apple Mac computer, use Google 

Chrome of Apple Safari.  Neptune Systems recommends the use Chrome in either case. 

1.

To access the Classic Dashboard of an APEX (any model), follow these steps:  a.

If using an Apple Mac computer, continue to Step 2.  b.

If using a Windows computer, type  http://apex  in your web browser’s address bar and press 

ENTER.    i.

If you get a 404 or Page Not Found error, continue to Step 2  ii.

If you get a login prompt that looks similar to this, skip to Step 4: 

2.

Use your display module to find the IP address of the APEX:  a.

Press the center button once  b.

Press the down arrow button 3 times to scroll to  System  

 

  c.

Press the center button once  d.

Press the down arrow button once to scroll to  Net Setup  

 

Page 7 

Version: 1.2 

  e.

Press the center button  f.

Press the down arrow button to scroll to  IP Address    g.

Press the center button once  h.

Note the IP address shown 

 

  i.

Press the center button once  j.

Press the up arrow button 3 times to scroll to  HTTP PORT  

  k.

Note the HTTP Port number shown.  In most cases, it will be “80”. 

3.

In your web browser’s address bar, enter the address of the APEX in this format:    http://ip‐address  

Do not enter any spaces.  For example, if the display shows  address in your browser address bar like this: 

, enter the IP 

 

Page 8 

  a.

If the HTTP port noted previously is something other than 80, add a semicolon and that port  number after the IP address in your address bar, like this: 

Version: 1.2 

b.

Press the ENTER key on your computer keyboard. 

4.

You should be presented with a logon prompt like this: 

 

 

5.

Enter the username and password for your Apex.   This might not the same username and password as  you use for APEX Fusion .  If you do not know the APEX’s username and password, try username ‘admin’  and password ‘1234’. 

6.

After login, you should now see a web page similar to this: 

7.

You have successfully accessed the Classic Dashboard.   Proceed to the next section. 

 

Page 9 

Version: 1.2 

 

   

CLONING YOUR APEX 

REVIEW THIS SECTION COMPLETELY BEFORE ACTUALLY DOING ANYTHING WITH THE 

NEW/REPLACEMENT APEX.     

 

DO NOT DISCONNECT ANY MODULES FROM THE OLD APEX UNTIL DIRECTED. 

DO NOT CONNECT ANY MODULES TO THE NEW APEX OR APPLY POWER TO THE NEW APEX 

UNTIL DIRECTED. 

 

DO NOT USE MOZILLA FIREFOX OR MICROSOFT EDGE BROWSERS DURING THE CLONING 

PROCESS      

Mozilla Firefox is fine to use for accessing an APEX’s Classic Dashboard in everyday use, but  during the APEX cloning process, the most recent version of Firefox will not work.  Microsoft 

Edge (included in Windows 10) also has known issues.  If you use a Windows computer, use 

Google Chrome or Microsoft Internet Explorer.  If you use an Apple Mac computer, use Google 

Chrome of Apple Safari.  Neptune Systems recommends the use Chrome in either case. 

 

NOTE : Throughout this procedure, “APEX” generally means all models of APEX‐family controllers.  

Where there are differences which apply only to Apex base units or only to APEX JR, the terms 

“APEX base unit” and “APEX JR” will be used to differentiate as needed. 

 

Create a configuration backup of the OLD APEX 

1.

 Do one of the following: 

If you can still access the  old  APEX’s Classic Dashboard, proceed to Step 2 

If you have a recently‐saved configuration backup file saved on your computer  AND  you cannot  access the Classic Dashboard of the OLD APEX, skip ahead to  Prepare the NEW APEX .    

2.

Using the Classic Dashboard, browse to the Configuration‐>Load/Save page 

3.

Click the  Save Apex Configuration  button 

 

4.

Save the file in the computer’s Downloads folder.  The default file name is ‘apexcfg.bin’.  It is  recommended that you change this name to include the current date, such as ‘apexcfg_02FEB2016.bin’  when saving the file to your computer. 

 

Page 10 

Version: 1.2 

5.

Got to Configuration‐>Misc Setup and make a screen shot of the Feed Interval settings, or jot them  down. 

6.

Got to Configuration‐>Network Setup and make a screen shot of the network settings, or jot them  down. 

7.

If you have customized the Season Table in the  old  Apex, make a screen shot of the Season Table page. 

 

 

Prepare the NEW APEX 

DO NOT DISCONNECT ANY MODULES FROM THE OLD APEX UNTIL DIRECTED. 

DO NOT CONNECT ANY MODULES TO THE NEW APEX UNTIL DIRECTED. 

 

1.

Remove the  new  APEX from the packaging. 

2.

Disconnect ONLY your display module from the  old  APEX and connect it to an AquaBus port on the  new  

APEX.   DO NOT CONNECT ANY OTHER MODULES TO THE NEW APEX YET, and DO NOT DISCONNECT 

ANY MODULES FROM THE OLD APEX OR FROM OTHER MODULES . 

3.

Disconnect the network cable from the  old  APEX and use that same cable to connect the  new  APEX to  your network. 

4.

Power up the new APEX or APEX JR 

For APEX base units: Connect the 12‐volt AC adapter to the POWER jack on the  new  APEX, and  plug the 12‐volt AC adapter into a regular electrical outlet. 

For APEX JR: Plug the APEX JR power cord into a regular electrical outlet. 

5.

Wait until the Status LED on the face of the  new  APEX is lit solid green. 

6.

Using the display menu, reset the  new  APEX to factory defaults.  a.

Press the center button once.  b.

Use the down button to scroll to  System .  c.

Press the center button once.  d.

Use the up button to scroll to  Init Memory.

  e.

Press the center button once.  f.

Scroll to  Init All  if needed (it should already be selected).  g.

Press the center button once.  h.

Press the  Yes  button when prompted “ Are you sure ?”.  i.

Wait until the Status LED on the face of the  new  APEX goes out then becomes lit solid green  again. 

7.

Upgrade the  new  APEX’s firmware and web pages to the latest version using the instructions and APEX  firmware files available at  https://www.neptunesystems.com/firmware/.

 

8.

Using the display menu, update the  new  APEX bootloader.  a.

Press the center button once.  b.

Press the down button 3 times to scroll to  System.

  c.

Press up button 2 times to scroll to  Bootloader Update.

  d.

Press the center button once.  e.

The display might go blank for several seconds, or might just flash momentarily. 

 

Page 11 

Version: 1.2 

f.

Wait until the screen returns, then use the down button to scroll to  Net Setup .  g.

Press the center button.  h.

Press the up button once to scroll to  Restart .  i.

The APEX will reboot; wait until the Status LED on the face of the APEX is lit solid green, then  proceed to the next section. 

 

 

 

Restore the Configuration Backup on the NEW APEX 

DO NOT DISCONNECT ANY MODULES FROM THE OLD APEX UNTIL DIRECTED. 

DO NOT CONNECT ANY MODULES TO THE NEW APEX UNTIL DIRECTED. 

 

1.

Access the Classic Dashboard of the  new  APEX.   Refer back to the  ACCESSING THE CLASSIC DASHBOARD   topic for instructions if needed. 

2.

Browse to the Configuration‐>Load/Save page. 

3.

Click the  Browse  button. 

 

4.

Locate the configuration backup file you saved previously.  The filename will be ‘apexcfg.bin’ unless you  saved it with a different name.   Select it, then click  Open  (or  OK ). 

 

Page 12 

Version: 1.2 

 

5.

Verify that the selected backup file name appears in dim gray text to the right of the  Browse  button. 

 

6.

Click the  Select All  button.  Verify that all options are selected, then de‐select the Network Configuration  option. 

 

Page 13 

Version: 1.2 

 

7.

Click the  Restore Configuration  button. 

 

8.

Watch the status LED on the face of the  new  APEX.  After a minute or two, the status LED will go out,  then will flash orange for several seconds, and finally be lit solid green. 

9.

In the browser tab or window for the  new  APEX, click the Dashboard menu. 

10.

Verify that tiles appear on the Classic Dashboard for your probes and outlets, and have the names they  had in the  old  APEX. 

11.

All outlets will be in AUTO.  If you had any outlets in the  old  APEX set to manual OFF or manual ON 

 

(instead of AUTO) as the normal state, check the outlet sliders using the Dashboard page and change  those outlets to the desired ON or OFF state. 

 

Enable REPLACEMENT MODE in the NEW APEX 

1.

Using the Classic Dashboard (Configuration‐>Clock Setup), or using the display menu (System‐>Clock 

Setup), update the  new  APEX clock settings for the proper time zone and Daylight Saving Time settings  for your geographic location. 

2.

Using the display module connected to the  new  APEX, enable REPLACEMENT MODE:  a.

Press the center button once.  b.

Press the down arrow 5 times to scroll to  Self Test.

  c.

Press the center button once.  d.

Press the up button twice. 

 

Page 14 

Version: 1.2 

e.

Press the down button twice.  f.

Press the center button once.  g.

The display screen should appear like this: 

 

  h.

Press the down button once to scroll to  Diag .  i.

Press the center button once.  j.

Press the down button twice to Replacement Mode, which usually will be  Off .

 (The word “Off” will be truncated”.)  k.

If you received a replacement APEX base unit under RMA, REPLACEMENT MODE might already  be enabled; if so, skip to Step 3.  l.

Press the center button once.  m.

The display will go blank briefly.  When it returns to normal, it should indicate that Replacement 

Mode is now  On.

 

3.

Using the display module connected to the  new  APEX, delete the display module:  a.

Press the Home button.  b.

Press the Center button once.  c.

Press the down button 2 times to scroll to  Setup.

  d.

Press the center button once.  e.

Press the down button 2 times to scroll to  Module   Setup.

  f.

Press the center button once.  g.

Press the down button 3 times to scroll to  Delete Module.

  h.

Press the center button once.  i.

A display ( Disp_#)  should be the only module listed and it should already be selected.  j.

Press the center button once.  k.

Press the  Yes  button, then disconnect the display module from the  new  APEX  immediately  after 

 appears on the display screen. 

4.

Wait 5‐10 seconds then:  a.

For APEX: disconnect the 12‐volt AC adapter and network cable from the  new  APEX.  b.

For APEX JR: unplug the  new  APEX JR power cord and disconnect the network cable. 

REPLACEMENT MODE will be automatically disabled at midnight (00:00) per the Apex’s clock.  

The next task must be completed before midnight of the day you enable REPLACEMENT 

MODE. 

 

 

 

 

Page 15 

Version: 1.2 

Replace the OLD APEX with the NEW APEX 

1.

Disconnect the Aquabus cable(s) from the  old  APEX base unit or APEX JR. 

2.

If using an APEX JR, disconnect device power cords from the APEX JR’s outlets, making note of which  power cord is plugged into which outlet. 

3.

Disconnect any Temp, pH, and/or ORP probes from the  old  APEX.  There is no need to disconnect probes  from any Probe Modules, leak sensors from an ALD, power cords from EnergyBars, etc. 

4.

Dismount the  old  APEX if necessary. 

5.

Mount the  new  APEX as desired. 

6.

Connect any Temp, pH, and/or ORP probes to the  new  APEX. 

7.

Connect the network cable to the  new  APEX. 

8.

Connect the AquaBus cables that you disconnected in Step 1 to the  new  APEX.   Do NOT reconnect the  display module yet . 

9.

If using an APEX JR, plug the APEX JR power cord into an electrical outlet. 

10.

Wait at least 60 seconds.   

11.

If you previously had a 12‐volt AC adapter connected to your  old  APEX base unit, reconnect it to the  new  APEX. 

12.

Connect the display module to any available AquaBus port of the  new  APEX system. 

13.

If using an APEX JR, plug device power cords into the same outlets previously used. 

 

Disable REPLACEMENT MODE 

1.

On the display module connected to the  new  APEX, disable REPLACEMENT MODE:  a.

Press the center button once.  b.

Press the down arrow 5 times to scroll to  Self Test.

  c.

Press the center button once.  d.

Press the up button twice.  e.

Press the down button twice.  f.

Press the center button once.  g.

Press the down button once to scroll to  Diag.

  h.

Press the center button once.  i.

Press the down button twice to scroll to  Replacement Mode.

  j.

If it shows as  On , press the center button to turn off.  The display will go blank briefly.  When it  returns to normal, it should indicate that  Replacement Mode  is now  Off.

 

2.

Press the Home button to exit the menu system. 

 

Additional Steps on the NEW APEX 

Calibrate the pH probe(s) on the  new  APEX base unit (not applicable for APEX JR). 

Using the Classic Dashboard, go to the Configuration‐>Misc Setup page.  o Due to a software bug in some firmware versions, the setting for Feed Interval D might not have  been transferred.  Check and update the Feed Interval D value if needed.  o Enable the Power Monitor feature  only  if you had used this feature with the  old  APEX. 

 

Page 16 

Version: 1.2 

Using the Classic Dashboard, go to the Configuration‐>Module Setup page.   Look at the table on the  lower section of the page.  If any modules appear with  (OLD)  in the SW column, update the firmware on  those modules using the upper part of that page.  For any modules that show OLD:  o Using the drop‐down menu on the upper left part of the Module Setup page, select a module which  shows (OLD).  o After the page reloads, click to select  Update Firmware   o Click  Submit Module Update   o Watch the space just above the bottom table to see the status of the module firmware update  process.  o Repeat for any other module needing to be updated. 

Verify that APEX Fusion is enabled, and update other Network Settings.  Using the Classic Dashboard, go  to the Configuration‐>Network Setup page.   Configure any non‐default network settings previously used  on your  old  Apex.  o Verify that the  Apex Fusion Enable  and  Apex Fusion Control Enable  options are both enabled.  Enable  both settings if they are disabled.  o If your old APEX had a user‐defined hostname, use the Classic Dashboard (Configuration‐>Network 

Setup) to change the  new  APEX’s hostname as desired.  o If you previously used a static IP address on your  old  Apex, you should now configure the  new  Apex  accordingly.  o If you previously used a non‐standard HTTP port on your  old  Apex, you should now configure the 

  new  Apex accordingly.  o Restart the APEX after any network settings are changed so these changes take effect. 

If you previously customized the APEX Season Table, recreate those modifications using the screen  shot obtained previously, or use websites such as  http://www.timeanddate.com/  and 

  http://aa.usno.navy.mil/data/docs/RS_OneYear.php

 to obtain current sunrise/sunset and lunar  cycle data. 

Link the NEW APEX with APEX Fusion 

1.

Sign in to APEX Fusion ( https://apexfusion.com

). 

2.

On the display module connected to the  new  APEX, press the center button once. 

3.

Press the down button to scroll to  ApexFusion: Link  then press the center button once. 

4.

If you do not see the word “Link” change to a 5‐character code after 5 seconds, press the center button  again. 

5.

The ApexFusion line should now show a 5‐character code which is the APEX Fusion token. 

6.

Click on your Fusion username in the upper right corner of the APEX Fusion web page. 

7.

Click                           

8.

Enter the 5‐character token from Step 5 in the Link Token field. 

9.

Click 

10.

 After the linking process completes, go to the Apex List page. 

11.

Locate your  old  APEX on the APEX List, which will probably have a         icon. 

12.

Click the         or         icon to the left of the  old  APEX, then click                     . 

13.

Enter your Fusion account password when prompted. 

 

Page 17 

Version: 1.2 