- Industrial & lab equipment
- Measuring, testing & control
- INFICON
- 580-015
- Original operating instructions
advertisement
Original operating instructions
IRwin S/SX/SXT/SXG/SXGT
Portable Methane Leak Detector
Catalog No.
580-000, 580-010, 580-015, 580-020, 580-030
From software version
3.01.01
mina66en1-15-(1904)
INFICON GmbH
Bonner Strasse 498
50968 Cologne, Germany
INFICON Table of Contents
Table of Contents
IRwin-Operating-instructions-mina66en1-15-(1904) iii
Table of Contents INFICON
iv IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Table of Contents
IRwin-Operating-instructions-mina66en1-15-(1904) v
Table of Contents INFICON
vi IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON About these instructions | 1
1 About these instructions
This document applies to the software version stated on the title page.
Product names may occur in the document, which are added for identification purposes only and belong to the respective owner of the rights.
1.1 Target groups
These Operating Instructions are written for trained and experienced gas pipeline survey operators.
1.2 Warnings
DANGER
Imminent hazard resulting in death or serious injuries
WARNING
Hazardous situation resulting in potential death or serious injuries
CAUTION
Hazardous situation resulting in minor injuries
NOTICE
Hazardous situation resulting in damage to property or the environment
IRwin-Operating-instructions-mina66en1-15-(1904) 7 / 128
2 | Safety INFICON
2 Safety
2.1 Intended use
The IRwin Methane Leak Detector is intended for professional use in residential and commercial areas. The operation of the equipment requires sufficient knowledge and experience in gas pipeline inspection. The different IRwin versions are equipped for
the measurement of different gases, see "Technical data [ } 22]", as well as for the
applications listed below:
• IRwin S
Above ground, Bar hole
• IRwin SX (EX certified)
Above ground, Bar hole, Confined Space, House, Gas Purity, Ex
• IRwin SXT (EX certified)
Above ground, Bar hole, Confined Space, House, Gas Purity, Ex, Ex Tox
• IRwin SXG (EX certified)
Above ground, Bar hole, Confined Space, House, Gas Purity, Ex, GC Ethane
Analysis
• IRwin SXGT (EX certified)
Above ground, Bar hole, Confined Space, House, Gas Purity, Ex, Ex Tox, GC
Ethane Analysis
The IRwin SX, IRwin SXG, IRwin SXT and IRwin SXGT are intrinsically safe and can be used in areas with potentially explosive atmospheres according to the EX rating
and EX application it is certified for (see "EX certification (intrinsic safety) [ } 24]").
These models are jointly referred to as IRwin SXnn in the following.
The IRwin S is not EX certified.
Pay attention to the safety instructions "Safe operation [ } 12]".
• You must assemble, operate and maintain the equipment only in compliance with these operating instructions.
• Use only within the allowed ambient conditions . Operating or storing the equipment outside the given range can result in erroneous readings and possible malfunction.
• Use the leak detector exclusively for the detection of the specified gases.
8 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON
Improper use • Do not suck liquids into the leak detector.
• Do not suck dirt or sand into the leak detector.
• Do not use the IRwin S in areas with potentially explosive atmospheres.
See also
2
2
EX certification (intrinsic safety) [ } 24]
2.2 Special conditions for safe use
Safety | 2
2.2.1 Intrinsic Safety (Ex protection)
IRwin is certified intrinsically safe to prevent ignition of flammable atmospheres.
IRwin holds an ATEX certificate for EU and corresponding certificates for several other
regions as shown by the attached certificates, see Certificates [ } 98]).
For details of rating, see EX certification (intrinsic safety) [ } 24].
The "X" after the respective type certificate number relates to the following special conditions for safe use:
The only device allowed to be connected to the charging socket on any of the models
• IRwin SX
• IRwin SXT
• IRwin SXG
• IRwin SXGT is the Ex Certified IRwin Charging adapter (PN 580-604).
The charger or car adapter is subsequently connected to the inlet of the Ex Certified
IRwin Charging Adapter.
This means that the charger for IRwin S must, under no circumstances, be connected directly to the charging socket on any of IRwin models SX, SXT, SXG, and SXGT. See
also Charging the battery [ } 35].
IRwin SXnn is certified as intrinsically safe at an ambient temperature of -20 - +50°C.
Further temperature restrictions apply to oxygen and toxic gas measurment of IRwin
SXT and SXGT. See Certified Gas Measurement.
IRwin SXnn is certified for use in Gas groups IIA, IIB and IIC and in Zones 0, 1 and 2.
The external surface of the enclosure is therefore slightly conductive to prevent electrostatic charging and sparks. Labels added for marking purposes must not be larger than 400 mm
2
. If larger the instrument should not be used in gas group IIC,
Zone 0 applications. Natural gas applications are not affected by this regulation.
IRwin-Operating-instructions-mina66en1-15-(1904) 9 / 128
2 | Safety INFICON
10 / 128
2.2.2 Certified Gas Measurement
IRwin SX, IRwin SXG, IRwin SXT, and IRwin SXGT are certified by TÜV Rheinland to
EN 60079-29-1 for measuring methane (CH
4
), propane C
3
H
8
, butane C
4
H
10
and natural gas concentration in units of percentage of lower flammability level. In IRwin we use the more widely known acronym LEL as synonymous with the more correct LFL.
IRwin SXT and IRwin SXGT are, in addition, also certified by TÜV Rheinland to
EN50104 and EN 45544 for measuring Oxygen (O
2
), Carbon Dioxide (CO
2
), Carbon
Monoxide (CO) and Hydrogen Sulfide (H
2
S).
The certification is valid in an ambient temperature of -15 to +40 °C and subject to
periodic calibration, see ”Calibration [ } 49]”.
IRwin S, IRwin SX and IRwin SXG are operable within -20 to +50 °C. The certification is valid in an ambient temperature of -15 to +40 °C. The accuracy may not be according to specification outside this range.
IRwin SXT and IRwin SXGT are operable within -20 to +40 °C. The certification is valid in an ambient temperature of -15 to +40 °C.
IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Safety | 2
2.3 Owner/supervisor responsibilities
Safety conscious operation
Staff qualifications
Non-authorized repairs forbidden
The following notes are for organizations or any person responsible for the safe and effective use of this equipment.
• Check that the leak detector is properly calibrated and not damaged before using it.
• Make sure you operate the leak detector in accordance with this manual.
• Adhere to the following regulations:
– Intended use
– General applicable safety and accident prevention regulations
– International, national and local standards and guidelines
– Additional equipment-related provisions and regulations
• Use only original parts or parts approved by the manufacturer.
• Keep this instruction manual available on site.
• Only properly trained staff should be permitted to work with and on the equipment.
The training must cover the actual equipment model used.
• Make sure that users have read and understood the operating instructions and all other applicable documents.
The manufacturer (INFICON) denies all responsibility for the compliance of this product with any of the type certificates for this product if any repairs or service involving opening the instrument enclosure (yellow box) has been performed by individuals or organizations not qualified therefore in writing by INFICON AB, Sweden.
The manufacturer (INFICON) denies all responsibility for the compliance of this product with any of the type certificates for this product if the equipment is used in any way not conforming to the instructions in this User Manual. Replacement of externally accessible parts such as probes and filters is allowed with the exception of the Hand
Probe and the hose between Hand Probe and IRwin detector. These two parts are certified anti-static and must not be replaced with non-original parts.
IRwin-Operating-instructions-mina66en1-15-(1904) 11 / 128
2 | Safety INFICON
2.4 Duties of the operator
• Read, observe, and follow the information in this manual and in the work instructions provided by the owner. This concerns in particular the safety instructions and warnings.
• Always observe the complete operating instructions for all work.
• If you have any questions about operation or maintenance that are not answered in this manual, please contact Customer Service.
2.5 Safe operation
Hazards due to chemical substances and explosive gases
Operation mode and
Ex
• Do not expose the leak detector to excess heat or a naked flame.
• Agressive substances such as solvents and other chemicals can damage the equipment.
• Adhere to the restrictions of use.
• Do not suck liquids into the leak detector. Proper maintenance of all filters reduces this risk.
The safety related modes Ex and Ex Tox has certified safety features. The detector monitors its functionality and proper function is indicated by an acoustic alive signal beeping every 20 s and by the LED being green. This requires that the leak detector is calibrated according to the instructions and intervals stated in this manual.
If the signal LED is red, this indicates an alarm or an error.
Green LED and short beep every 20 s: Function OK.
Red LED: Gas alarm or instrument error.
Yellow LED: Special state Special state indicates functioning system but alarms are typically not active. This special state may occur for example during calibration or when changing a setting in the leak detector.
IRwin Methane Leak Detector was built according to the state-of-the-art and the recognized safety regulations. Nevertheless, improper use may result in risk to life and limb on the part of the user or third parties, or damage to the leak detector or other property may occur.
12 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Scope of delivery | 3
3 Scope of delivery
The following is included in the delivery of IRwin:
Article (catalog number)
Harness (580-405)
Charger for IRwin S, 100-240V (580-603)
Car adapter for IRwin S, 12V (580-602)
Charger cable for car (591-361)
Charger for IRwin SXnn models, 100-240V (580-605)
Certified IRwin Charging Adapter (580-604)
Operating instructions
*Only valid for IRwin S up to serial number 92000700.
Quantity
1
1 (IRwin S only)*
1
1
1 (IRwin S only)*
1
1
IRwin Accessory Kit (580-712):
Article (catalog number)
Rod 850mm (580-150)
Extension Rod 150mm (580-160)
Carpet (580-211)
Bell (580-301)
Carrying Case IRwin (580-450)
Filter-kit (580-700)
Bar Hole Probe (580-115)
Extension Connector (580-220)
► Check that the delivery is complete.
Quantity
1
1
1
1
1
1
1
1
IRwin-Operating-instructions-mina66en1-15-(1904) 13 / 128
4 | Description INFICON
4 Description
4.1 Function
INFICON single-cell wide range technology
(patent pending)
The IRwin is a portable methane leak detector for gas pipeline survey. It measures following gases depending on the leak detector version:
• Methane
• Carbon dioxide
• Ethane
• Propane
• Butane
• Oxygen
• Hydrogen sulfide
• Carbon monoxide
There are several probes available for the leak detector:
• The carpet probe is used for surveying paved or smooth surfaces. Pull or push the carpet along the surface.
• Use the bell probe for measuring gas at specific points. Press the bell firmly down to pull gas through the surface.
• The hand probe is used for checking features on or in buildings. The hand probe also constitutes the handle for the carpet probe and bell probe.
• The Bar Hole probe is used to sample air from drilled holes. Push into hole and tighten screw to expand rubber before connecting sample line.
• The optional Hand Probe Flexible Extension can be connected to the hand probe or rod for easier access.
• The Swan Neck Probe is used on uneven surfaces and in bushes etc.
• The Flexi Bell is used for simple survey including meters and piping. The bell adapts to pipe geometry.
IRwin methane leak detector uses the unique INFICON single-cell wide range technology for detecting all concentrations from 1 ppm to 100 Vol% methane. This technology is very sensitive to the ethane content of natural gas. Therefore, large deviations can occur at higher methane concentrations if the ethane concentration is not compensated for. The instrument has built-in routines for ethane compensation.
The actual ethane concentration can be determined using the IR Ethane Analysis function. This can be used to automatically set the compensation concentration. If preferred you can instead set the ethane content manually. The safety certified modes
14 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Description | 4
IRwin SXG and SXGT models also have GC Ethane Analysis capability for distinguishing between Natural gas (NG), liquid petroleum gas (LPG) and swamp gas by determining the presence of methane, ethane and propane in the sampled gas.
The GC Ethane Analysis function is not affecting the ethane compensation.
4.2 Design of the leak detector
3
4
5
1
Fig. 1: Design of the Leak Detector
1
2
3
Reference inlet
Display
Gas inlet
2
4
5
Gas outlet
Charging socket
IRwin-Operating-instructions-mina66en1-15-(1904) 15 / 128
4 | Description INFICON
Carpet probe
4.3 Probes
WARNING
Risk of explosion due to not approved probe parts.
The original INFICON Hand probe is the only part of the probe system that is certified for use in potentially explosive areas. Other parts of the probe system, or third party accessories are not allowed in potentially explosive areas.
► Pay proper attention to safety when using other accessories for locating and pinpointing gas leaks in none classified areas.
When you are looking for gas leaks on paved surfaces, use the Carpet probe. Pull or push the Carpet probe along the surface.
Bell probe
Fig. 2: Carpet probe
When searching for or measuring gas leaks at particular spots, use the Bell probe.
This can be used to pull gas through most pavings. It is also suitable for emergency testing if surface is wet.
Fig. 3: Bell probe
16 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON
Hand probe
Description | 4
You can use the Hand probe to search for gas in building installations. The hand probe also constitutes the handle for the Carpet probe and for the Bell probe.
Bar Hole Probe
Fig. 4: Hand probe
The Bar Hole Probe is used to pump out and take samples from holes drilled in the asphalt or concrete when pin-pointing a leak. Note that the Bell Probe can be used first to reduce the number of holes.
Hand Probe Flexible
Extension
Fig. 5: Bar Hole Probe
The flexible extension is practical when locating leaks in hard to reach installations.
Use the Extension Connector to attach it to the long rod to reach even further.
Fig. 6: Hand Probe Flexible Extension
IRwin-Operating-instructions-mina66en1-15-(1904) 17 / 128
4 | Description
Swan Neck
INFICON
The Swan Neck is used on uneven surfaces where the carpet probe is unpractical.
Can also be practical in bushes and other “confined” areas.
Flexi Bell
Fig. 7: Swan Neck
The Flexi Bell is practical for surveys including frequent entries into gardens and check of meters and piping. The softer bell adapts to pipe geometry.
Fig. 8: Flexi Bell
18 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Description | 4
4.4 Display
1 1
6
3
1
2
5
9 7
(10 PPM)
1 2
7
9
8
5
Status indicator LED
Buzzer
Above ground
Fig. 9: Display IRwin
1
3
5
Navigation buttons
Battery status indicator
Measured value, target gas and unit
2
4
Navigation and on / off button
Mode icon
6 Measurement range (in % or PPM
CH
4
depending on the operation mode)
8 Operation mode and menu bar
10 PPM threshold in Above Ground and Confined Space modes.
Navigation buttons
Menu
Settings
Operation
Information
Diagnosis
Table 1: Navigation buttons
Upper left
Upper left after pressing “Menu”
Upper right
Lower left after pressing “Menu”
At lower right after pressing “Menu”
To select a menu, press the navigation button to go to desired tab or field.
Press check mark button , select or open a highlighted option.
IRwin-Operating-instructions-mina66en1-15-(1904) 19 / 128
4 | Description INFICON
Function buttons OK/Select/Confirm
Navigation buttons (in this case go right)
Cancel process
Start process
Start data logging
Stop data logging
Mute the alarm
Table 2: Function buttons
Symbols of the status bar
Mode icons
Bluetooth activated
Symbol ON = GPS connected and tracking
Symbol flashing = GPS not connected
Battery status indicator
The color of the battery status indicator shows the remaining runtime before shutdown:
White: > 1h remaining
Orange: < 1h remaining
Red flashing: < 10 min remaining
Progress bars for calibration processes etc.
White bar: Process has not yet started
Light green bar: Process running
Dark green bar: Process completed
Red bar: Process failed
Table 3: Status indicators
•
•
Operation mode- Above ground
Operation mode- Bar hole
• Operation mode- Confined Space
• Operation mode- House
• Operation mode- Gas Purity
• Operation mode- Ex
• Operation mode- Ex Tox
• Operating mode- GC Ethane Analysis
• Operating mode- Universal
20 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON
Permission levels
Status LED
Description | 4
Symbol
Table 4: Permission levels
Green
Permission level
Basic
Intermediate
Advanced (Supervisor)
Password none
1111
1422
Cyan
Blue
Red
Yellow
• Leak detector is charging normal
(power off)
• Normal operation in the modes Ex and
Ex Tox.
Leak detector is charging fast.
Normal operation
(in all operation modes except Ex and Ex
Tox)
• Alarm
• Error message
The leak detector is in a state in which functions like for example alarms could be disabled. For example, during warm up or calibration.
IRwin-Operating-instructions-mina66en1-15-(1904) 21 / 128
4 | Description INFICON
4.5 Technical data
Name
Weight
Dimensions
(lxwxh)
Mechanical data
580-000
IRwin S
1.4 kg (3 lb.)
197 x 256 x 62 mm (6.6 x 9.7 x
2.3 in.)
580-010
IRwin SX
1.6 kg (3.5 lb.)
197 x 256 x 62 mm (6.6 x 9.7 x
2.3 in.)
580-015
IRwin SXT
1.6 kg (3.5 lb.)
197 x 256 x 62 mm (6.6 x 9.7 x
2.3 in.)
580-020
IRwin SXG
1.6 kg (3.5 lb.)
197 x 256 x 62 mm (6.6 x 9.7 x
2.3 in.)
580-030
IRwin SXGT
1.6 kg (3.5 lb.)
197 x 256 x 62 mm (6.6 x 9.7 x
2.3 in.)
Electrical data
Name
Power input
Memory capacity for measured data
Type of protection
Electronic interfaces
580-000
IRwin S
4 A
64 MB
IP54
Bluetooth 3.0
Class 1
Battery operating time
8 h as verified during certification, typical value 9 h
Battery charging time
4 h from empty to full charge (3 h for fast charging). 4 hours operation by 1 hour fast loading
Battery Lithium-Ion,
580-010
IRwin SX
4 A
64 MB
IP54
Bluetooth 3.0
Class 1
8 h as verified during certification, typical value 9 h
4 h from empty to full charge (3 h for fast charging). 4 hours operation by 1 hour fast loading
Lithium-Ion
580-015
IRwin SXT
4 A
64 MB
IP54
Bluetooth 3.0
Class 1
8 h as verified during certification, typical value 9 h
4 h from empty to full charge (3 h for fast charging). 4 hours operation by 1 hour fast loading
Lithium-Ion
580-020
IRwin SXG
4 A
64 MB
IP54
Bluetooth 3.0
Class 1
8 h as verified during certification, typical value 9 h
4 h from empty to full charge (3 h for fast charging). 4 hours operation by 1 hour fast loading
Lithium-Ion
580-030
IRwin SXGT
4 A
64 MB
IP54
Bluetooth 3.0
Class 1
8 h as verified during certification, typical value 9 h
4 h from empty to full charge (3 h for fast charging). 4 hours operation by 1 hour fast loading
Lithium-Ion
10.0 Ah / 7.2 V /
73 Wh
10.0 Ah / 7.2 V /
73 Wh
10.0 Ah / 7.2 V /
73 Wh
10.0 Ah / 7.2 V /
73 Wh
10.0 Ah / 7.2 V /
73 Wh
22 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Description | 4
Physical data
Name
Noise level
580-000
IRwin S
< 70 dB (A)
580-010
IRwin SX
< 70 dB (A)
580-015
IRwin SXT
< 70 dB (A)
580-020
IRwin SXG
< 70 dB (A)
580-030
IRwin SXGT
< 70 dB (A)
Detectable gases
Methane, ethane, propane, butane, carbon dioxide
Methane, carbon dioxide, ethane, propane, butane
Methane, carbon dioxide, ethane, propane, butane, oxygen, hydrogen sulfide, carbon monoxide
Methane, carbon dioxide, ethane, propane, butane
Methane, carbon dioxide, ethane, propane, butane, oxygen, hydrogen sulfide, carbon monoxide
Gas flow through sniffer line
Typically 60 l/h Typically 60 l/h Typically 60 l/h Typically 60 l/h Typically 60 l/h
Ambient conditions
Name
Max. altitude above sea level
Permissible ambient temperature
(during operation)
580-000
IRwin S
2000 m
-20° - +50°C (-4
- 122°F)
Storage temperature
-25 - +70°C (-13
- 158°F)
Pressure range 80 kPa - 120 kPa
Max. relative humidity
95% (noncondensing)
580-010
IRwin SX
2000 m
-20 - 50°C (-4 -
122°F)
-25 - 70°C (-13 -
158°F)
80 kPa - 120 kPa
95% (noncondensing)
580-015
IRwin SXT
2000 m
-15 - 40°C (-5 -
104°F)
-25 - 70°C (-13 -
158°F)
80 kPa - 120 kPa
95% (noncondensing)
580-020
IRwin SXG
2000 m
-20 - 50°C (-4 -
122°F)
580-030
IRwin SXGT
2000 m
-25 - 70°C (-13 -
158°F)
-25 - 70°C (-13 -
158°F)
80 kPa - 120 kPa
95% (noncondensing)
-15 - 40°C (-5 -
104°F)
80 kPa - 120 kPa
95% (noncondensing)
IRwin-Operating-instructions-mina66en1-15-(1904) 23 / 128
4 | Description INFICON
Type certificate
4.5.1 EX certification (intrinsic safety)
Type certificates, see "Certificates [ } 98]"
The products:
• IRwin SX 580-010,
• IRwin SXT 580-015,
• IRwin SXG580-020,
• IRwin SXGT580-030, are Ex certified with rating according to the following table. The IRwin S 580-000 is not
Ex certified.
EX rating
EX rating
EX ranges of applications
Ex ia IIC T3, T amb
: -20°C - +50°C,
II 1G (EPL Ga)
USA: Intrinsically safe, Class I, Division 1, Groups A,
B, C and D.
• Zones: 0, 1 and 2
• Gas groups: IIA, IIB and IIC
• Temperature classes: T1, T2 and T3
• USA: Zone 0, Class I, Division 1, Groups A, B, C and D.
4.5.2 Sensors
Basic sensor data
Sensor performance data
Gas (measurement range)
CH
4
, C
3
H
8
, C
CO
2
O
2
, CO, H
2
S
4
H
10
Sensor
Infrared (IR)
Infrared (IR)
Electrochemical
Warm up time
< 30 s
< 30 s
< 120 s
O
2
sensor: 1 - 12 hours if stored with flat battery.
The calibration gases defined below are the typical gases and gas mixes used to calibrate the different modes. For optimum performance adjust these settings to the
actual concentrations of the gases used. See Calibration [ } 49].
24 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Description | 4
Universal mode
This mode is available if Universal Mode option is enabled. See Operating mode
Measurement principle
Measurement range
Resolution
Infrared (IR)
CH
4
, C
3
H
8
, C
4
H
10
: 1 ppm - 100 Vol%
The following is also displayed when total NG is above 2.7 Vol%:
C
2
H
6
content in NG: 0.5 - 8 Vol%
CO
2
: 0.1 - 20 Vol%
CH
4
: 1 ppm (0 - 100 Vol%), 10 ppm (110 - 990 ppm), 0.1 Vol%
(0.1 - 2.7 Vol%), 1 Vol% (<2.7 Vol%)
Measurement error
CO
2
: 0.1 Vol% (0.1 - 1 Vol%), 1 Vol% (<1 Vol%)
C
2
H
6
: 0.1 Vol% (Auto test) 0.5 Vol% (manual selection)
CH
4
: +5/-2 ppm (<20 ppm) ±10 Vol% (20 - 50 ppm) ±5 % (50 ppm - 2.2 Vol%) ±20 % (2.2 - 2.7 Vol%), ±3 Vol% (>2.7 Vol%)
Response time
Recovery time
CO
2
: ±1% (<10 Vol%), ±20% (>10 Vol%) t
50
<3 s (<2.7 Vol%), <5 s (>2.7 Vol%), t
90
<5 s (<2.7 Vol%), <10 s (>2.7 Vol%) t
50
<3 s (<2.7 Vol%), <5 s (>2.7 Vol%), t
10
<5 s (<2.7 Vol%), <10 s (>2.7 Vol%)
Cross sensitivity All hydrocarbons CxHy
Lifetime
Calibration gases
CO
2
and H
2
O: Negligible
Guaranteed 1 year, expected > 2 years
Fresh air, CH
4
in synthetic air: 10 ppm, 100 ppm 2.2 Vol%, 100
Vol% CO
2
, 20 Vol%
Note: IRwin allows other concentrations and reduced number of gases. The measurement error may fall outside specification if choosing other set-up than the recommended.
IRwin-Operating-instructions-mina66en1-15-(1904) 25 / 128
4 | Description INFICON
Above ground mode:
This mode is not available if Universal mode option is activated.
Measurement principle
Infrared (IR)
Measurement range
Resolution
CH
4
, C
3
H
8
, C
4
H
10
: 1 ppm - 5 Vol%
Calibration gases
1 ppm (0 - 100 ppm), 10 ppm (110 - 990 ppm), 0,1 Vol% (0,1 - 5
Vol%)
Measurement error
+5/-2 ppm (<20 ppm) ±10% (20 - 50 ppm) ±5% (50 ppm - 2.2
Vol%) ±20% (2.2 - 5 Vol%)
Response time
Recovery time t
50
<3 s, t
90
<5 s t
50
<3 s, t
10
<5 s
Cross sensitivity All hydrocarbons CxHy
Lifetime Guaranteed 1 year, expected > 2 years
Fresh air, CH
4
in synthetic air: 10 ppm, 100 ppm, 2.2 Vol%
Bar hole mode:
This mode is not available if Universal mode option is activated.
Measurement principle
Infrared (IR)
Measurement range
CH
4
, C
3
H
8
, C
4
H
10
: 0.5 - 100 Vol%
C
2
H
6
content in CNG: 0.5 - 8 Vol%
Resolution
CO
2
: 0.1 - 20 Vol%
CH
4
& CO
2
: 0.1 Vol% (0.1 - 1 Vol%), 1 Vol% (>1 Vol%)
Measurement error
C
2
H
6
: 0.1 Vol% (Auto test) 0.5 Vol% (manual selection)
CH
4
: ±3 Vol%
Response time
Recovery time
CO
2
: ±1% (<10 Vol%), ±20% (>10 Vol%) t
50
<5 s, t
90
<10 s t
50
<5 s, t
10
<10 s
Cross sensitivity All hydrocarbons CxHy
Lifetime
Calibration gases
CO
2
: Negligible
Guaranteed 1 year, expected > 2 years
Fresh air, CH
4
in synthetic air: 2.2 Vol%, 100 Vol%. CO
2
, 20 Vol
%
26 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Description | 4
GC Ethane Analysis mode:
Measurement principle
Measurement range
Capability
Cycle time*:
Gas Chromatography separation with semiconductor detector
1000 ppm gas needed in sample. This refers to total hydrocarbon content (i.e., natural gas, swamp gas or LPG etc.).
Can identify Natural Gas at 0.5% C
2
H
6 level.
Automatic interpretation of result as either of:
• NG with Ethane detected
• Methane detected
• LPG detected
• Gas type not identified (given if result is not clear)
Ethane only: 2 min at 25°C, 3 min at 0°C
Ethane and Propane: 5 min at 25°C, Propane detection not possible at 0°C
3 min Purge time
(before new GC test):
Lifetime Separation column: Guaranteed 3 years, expected > 10 years
Semiconductor sensor: Guaranteed 1 year, expected 3 years.
* Times given are valid after 1 h operation in given ambient temperature. Times include 30 s sensor purge before analysis. Three minutes column purge after analysis is not included. The column purge can be postponed.
IRwin-Operating-instructions-mina66en1-15-(1904) 27 / 128
4 | Description INFICON
House mode:
This mode is not available if Universal mode option is activated.
Measurement principle
Infrared (IR)
Measurement range
Resolution
CH
4
, C
3
H
8
, C
4
H
10
: 1 ppm - 5 Vol%
Calibration gases
1 ppm (0 - 100 ppm), 10 ppm (110 - 990 ppm), 0.1 Vol% (0.1 - 1
Vol%), 1 Vol% (>1 Vol%)
Measurement error
+5/-2 ppm (<20 ppm) ±10% (20 - 50 ppm) ±5% (50 ppm - 2.2
Vol%) ±20% (2.2 - 5 Vol%)
Response time
Recovery time t
50
<3 s, t
90
<5 s t
50
<3 s, t
10
<5 s
Cross sensitivity All hydrocarbons CxHy
Lifetime Guaranteed 1 year, expected > 2 years
Fresh air, CH
4
in synthetic air: 10 ppm, 100 ppm, 2.2 Vol%
28 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Description | 4
Confined Space mode:
This mode is not available if Universal mode option is activated.
Measurement principle
Measurement range
Resolution
Infrared (IR)
CH
4
, 1 ppm – 100 Vol% for natural gas with up to 8% C2H6
(ethane) content
1 ppm (0 - 100 ppm), 10 ppm (110 - 990 ppm), 0.1 Vol% (0.1 - 1
Vol%), 1 Vol% (>1 Vol%)
Measurement error
+5/-2 ppm (<20 ppm) ±10% (20 - 50 ppm) ±5% (50 ppm - 2.2
Vol%) ±3 Vol% (2.2 - 100 Vol%)
Response time
Recovery time t
50
<5 s, t
90
<10 s t
50
<5 s, t
10
<10 s
Cross sensitivity All hydrocarbons CxHy
Lifetime Guaranteed 1 year, expected > 2 years
Calibration gas Fresh air, CH
4
in synthetic air: 10 ppm, 100 ppm, 2.2 Vol%, 100
Vol%
Gas Purity mode:
This mode is not available if Universal mode option is activated.
Measurement principle
Measurement range
Resolution
Infrared (IR)
CH
4
: 0.5 - 100 Vol%
0.5 Vol% (0.5 - 1 Vol%), 1 Vol% (>1 Vol%)
Measurement ±3 Vol%
Response time
Recovery time t
50
<5 s, t
90
<10 s t
50
<5 s, t
10
<10 s
Cross sensitivity All hydrocarbons CxHy
Lifetime Guaranteed 1 year, expected > 2 years
Calibration gas Fresh air, CH
4
in synthetic air: 2.2 Vol%, 100 Vol%
IRwin-Operating-instructions-mina66en1-15-(1904) 29 / 128
4 | Description
30 / 128
INFICON
Ex mode:
This mode is not available in IRwin S.
Measurement principle
Measurement range
Resolution
Measurement error
Infrared (IR)
CH
4
, C
1%LEL
3
H
±5%LEL
8
, C
4
H
10
: 1 - 100% LEL
Response time
Recovery time
CH
4
: t
50
= 3 s, t
90
= 5 s t
50
<3 s, t
10
<5 s
Cross sensitivity All hydrocarbons CxHy
Lifetime
Calibration gas
Guaranteed 1 year, expected > 2 years
Fresh air, CH
4
in synthetic air: 2.2 Vol%
Ex Tox mode:
This mode is only available in IRwin SXT and SXGT.
Measurement principle
Measurement range
Infrared (IR)
Resolution
Measurement error
Response time
CH
4
, C
3
H
8
, C
4
H
10
: 1 - 100% LEL
CO
2
: 0.1 - 5 Vol%
O
2
: 0.1 - 25 Vol%
CO: 1 - 500 ppm
H
2
S: 1 - 400 ppm
CH
4
, C
3
H
8
, C
4
H
10
: 1% LEL
CO
2
: 0.05 Vol%
O
2
: 0.1 Vol%
CO: 1 ppm
H
2
S: 0.1 ppm
CH
4
: ±5 % LEL C
3
H
8
: ±40% of indicated value C
4
H
10
: ±50% of indicated value
CO
2
: ±0.1 Vol%
O
2
: ±0.3 Vol%
CO: ±3 ppm
H
2
S: ±0.5 ppm
CH
4
: t
50
<3 s, t
90
<5 s
IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Description | 4
Recovery time
Sensitivity drift
O
2
: t
50
<16 s, t
90
<40 s
CO: t
50
< 15 s, t
90
<30 s
CO
2
: t
50
<12 s, t
90
<36 s
H
2
S: t
50
<18 s, t
90
<66 s
CH
4
: t
50
<3 s, t
10
<5 s
CO
2
: t
50
<12 s, t
10
<36 s
O
2
: t
50
<16 s, t
10
<39 s
CO: t
50
<15 s, t
10
<30 s
H
2
S: t
50
< 18 s, t
10
<66 s
CH
4
, C
3
H
8
, C
4
H
10
: <±5 %LEL in 1 month
CO
2
: <±5 Vol% in 1 month
O
2
: <±1 Vol% in 3 months
CO: <±4% in 12 months
H
2
S: <±2% in 12 months
Cross sensitivity CH
4
: All hydrocarbons C x
H y
.
O
2
: negligible
CO: <12% of applied H
2
S concentration, <8% of applied H
2 concentration
H
2
S: <1.5% of applied CO concentration, <0.3% of applied H
2 concentration
Lifetime
Calibration gas
CO
2
: negligible
CH
4
, C
3
H
8
, C
4
H
10
: Guaranteed 1 year, expected > 2 years
CO
2
: Guaranteed 1 year, expected > 2 years
O
2
: Guaranteed 4 years, expected > 5 years
CO: Guaranteed 2 years, expected > 3 years
H
2
S: Guaranteed 2 years, expected > 3 years
Fresh air, ToxMix (CH
4
2.2 Vol%, CO
2
2 Vol%, CO 40 ppm, H
2
S
40 ppm, O
2
0 Vol% in N
2
)
IRwin-Operating-instructions-mina66en1-15-(1904) 31 / 128
4 | Description
32 / 128
INFICON
4.6 Factory settings
Parameter
Screen timeout (auto standby)
Factory Setting
30 s
C
4
H
10
AL1
C
4
H
10
AL2
C
4
H
10
AL3
CO
2
AL1
CO
2
AL2
CO
2
AL3
CO
2
STEL
CO
2
LTEL
CO AL1
CO AL2
CO AL3
CO STEL
CO LTEL
H
2
S AL1
Brightness
Select Gas (Ex)
CH
4
PPM Alarm
10
Screensaver (IRwin S only) enabled
PPM Adjustment Factor 1.0
CH
4
10 PPM
CH
CH
CH
C
3
4
4
4
H
AL1
AL2
AL3
8
PPM alarm
C
3
H
8
AL1
C
3
H
8
AL2
C
3
H
8
AL3
C
4
H
10
PPM alarm
10% LEL
50% LEL
100% LEL
10 PPM
10% LEL
50% LEL
100% LEL
10 PPM
10% LEL
50% LEL
100% LEL
0.5 Vol%
1.0 Vol%
5 Vol%
1.0 Vol%
0.5 Vol%
30 ppm
60 ppm
500 ppm
30 ppm
30 ppm
10 ppm
IRwin-Operating-instructions-mina66en1-15-(1904)
Options
Off
5, 30 s
1, 2, 5, 10, 20, 30 min
1, 2
1 -10 enabled or disabled
1.0, 1.2, 1.4, 1.6, 1.8, 2.0
CH
4
, C
3
H
8
, C
4
H
10
3, 5, 10, 15, 20, 25, 50,
100
3, 5, 10
30, 40, 50
80, 90, 100
3, 5, 10, 15, 20, 25, 50,
100
3, 5, 10
30, 40, 50
80, 90, 100
3, 5, 10, 15, 20, 25, 50,
100
3, 5, 10
30, 40, 50
80, 90, 100
0.1, 0.2, 0.3, 0.4, 0.5
0.6, 0.7, 0.8, 0.9, 1.0
1.0, 2.0, 3.0, 4.0, 5.0
0.5, 0.6, 0.7, 0.8, 0.9, 1.0
0.1, 0.2, 0.3, 0.4, 0.5
10, 20, 30
40, 50, 60
100, 200, 300, 400, 500
10, 20, 30
10, 20, 30
3, 5, 7, 10
INFICON
Parameter
H
2
S AL2
H
2
S AL3
H
2
S STEL
H
2
S LTEL
O
2
AL1
O
2
AL2
O
2
AL3
Calibration reminder
Function test reminder
100% CH
4
LEL conc
Deadband Suppression
Universal Mode
AbG Cal Setup
AbG Fnc Test conc
Time
Language
Factory Setting
20 ppm
100 ppm
10 ppm
10 ppm
10 Vol%
18 Vol%
23 Vol%
Off
Off
4.4 Vol%
Disabled
Disabled
Classic
10 ppm
UTC +2 h.
English
Ethane % 0%
Log to file
Logging Interval (log file period)
Start Mode enabled
3 s
Default (Ex or Above
Ground for IRwin S)
Auto-rotate Screen
Re-test C
2
H
6
in Bar Hole
Measuring duration (Auto bar hole)
Evacuation duration (Auto bar hole)
CO
2
-limit at evacuation
(Auto bar hole)
Off
Off
10 s
3 min
Off
Description | 4
Options
10, 15, 20
50, 60, 70, 80, 90, 100
3, 5, 7, 10
3, 5, 7, 10
3, 5, 10, 15
16, 17, 18, 19, 20
21, 22, 23, 24, 25
Off, 1 - 7, 14, 30 days
Off, 2, 4, 8 h.
4.4 Vol%, 5.0 Vol%
Disabled/ Enabled
Disabled/ Enabled
Generic / Classic
10 ppm, 500 ppm, 2.5%
-11 - +12 h.
English, Deutsch, Italiano,
Nederlands, Polskie,
Chinese
0 - 8% (in increments of
0.5%) enabled / disabled
1, 2, 3, 5, 10, 30 s, 1 min
Default, Last Used, Ex, Ex
Tox, Universal (if enabled),
GC Ethane Analysis
On, Off
Off, startup, always
10, 15, 20, 25, 30 s
3, 4, 5, 10, 15 min
Off, 1, 2, 3, 4, 5%
IRwin-Operating-instructions-mina66en1-15-(1904) 33 / 128
4 | Description
4.7 Concentration of calibration and test gases
Calibration routine
Ex (2.x% CH
4
)
Above Ground and
Universal (low range)
Bar Hole CH
4
and
Universal (high range)
GC Ethane Analysis
Option 1 mix*
GC Ethane Analysis
Option 2 mix*
Default gas
2.2 Vol% CH
4
10 ppm CH
4
100 ppm CH
4
2.2 Vol% CH
4
2.2 Vol% CH
4
100 Vol% CH
4
Bar Hole CO
2
and
Universal CO
2
20.0 Vol% CO
Ex Tox (Tox) Mixture of: 2.2 Vol% CH
4
2
2.0 Vol% CO
2
0 Vol% O
2
40 ppm H
2
S
40 ppm CO
Balance: N
2
1 Vol% CH
4
50 ppm C
2
H
6
1000 ppm C
3
H
8
1 Vol% CH
4
50 ppm C
2
H
6
Adjustable range
2.0 - 2.5 Vol% in 0.1
Vol% increments
Info
Select gas close to 50% LEL
8 - 15 ppm in 0.1 ppm increments
All three needed in "Classic" AbG
Cal setup.
80 - 1100 ppm in 1 ppm increments
1.5 -3.0 Vol% in 0.1
Vol% increments
Select one, two or three in
"Generic" setup.
Select gas close to 50% LEL 2.0 - 2.5 Vol% in 0.1
Vol% increments
80 - 100 Vol% in 1
Vol% increments
10.0 - 20.0 Vol% in
0.1 Vol% increments
1.4 - 2.7 Vol% in 0.1
Vol% increments
Select gas close to 50% LEL
0.5 - 3.0 Vol% in 0.1
Vol% increments
0.0 - 18.0 Vol% in 0.1
Vol% increments
4 - 50 ppm in 1 ppm increments
20 - 160 ppm in 1 ppm increments
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Timing adjustment only
Timing adjustment only
Timing adjustment only
Timing adjustment only
Timing adjustment only
*Used only for Function Test of GC Ethane Analysis.
Balance gas is Synthetic Air if nothing else stated (Ex Tox balance is N
2
).
INFICON
34 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Getting started | 5
5 Getting started
5.1 Charging the battery
WARNING
Explosion hazard
The equipment for charging IRwin is not ATEX certified/ex-protected.
► Never charge IRwin (all models) in potentially explosive atmospheres.
WARNING
Using the wrong charger could damage the explosion protection of IRwin
SXnn models.
The only equipment allowed to be connected to the Charging socket of any of the
IRwin SXnn models is the Ex Certified IRwin Charging Adapter. Connect this adapter to the leak detector first and then connect the charger or car cable to the adapter.
NOTICE
Fast charging reduces the lifetime of the battery.
► Do not use fast charging regularly.
► The ambient temperature during the charging process should be between 10 and
30 °C.
IRwin-Operating-instructions-mina66en1-15-(1904) 35 / 128
5 | Getting started INFICON
Accessories for charging for IRwin S up to serial number 929000704
Charger for IRwin S, 100-240V
(580-603) or
Car adapter for IRwin S, 12V
(580-602)
Charge IRwin S
1 Switch off IRwin S.
2 Connect the "Charger for IRwin S, 100-240V (580-603)" or "Car adapter for
IRwin S, 12V (580-602)" to the charging inlet of IRwin S. Align the red marks of the charging socket and charging plug with each other.
ð The status LED is green when the battery charges normal.
Fast charging IRwin S
1 For charging the leak detector faster switch on the IRwin S.
2 Connect the "Charger for IRwin S, 100-240V (580-603)" or "Car adapter for
IRwin S, 12V (580-602)" to the charging inlet of IRwin S.
3 Select fast charge when the pop-up is shown on the display.
ð The signal LED is turquoise when the battery is charging at a fast rate.
36 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Getting started | 5
Accessories for charging for IRwin SXnn and IRwin S from serial number 929000705
or
Charger cable for car (591-361)
Charger for IRwin SXnn models,
100-240V (580-605)
Certified IRwin Charging Adapter
(580-604)
Charging IRwin SXnn and IRwin S from serial number 929000705
1 Switch off IRwin SXnn.
2 Connect the “Certified IRwin Charging Adapter (580-604)“ to the charging inlet of IRwin SXnn. Align the red markings of the charging socket and charging plug to each other.
3 Connect the "Charger for IRwin SXnn models, 100-240V (580-605)" or "Charger cable for car (591-361)" to the “Certified IRwin Charging Adapter (580-604)".
ð The status LED is green when battery is charging normally. The status LED goes out when charging is completed.
IRwin-Operating-instructions-mina66en1-15-(1904) 37 / 128
5 | Getting started INFICON
Fast charging IRwin SXnn
► When the signal LED is green and indicating normal charge, press the upper right button.
ð The detector is now charging fast and the status LED turns cyan.
100%
Short charge
Normal
4 h from empty to full charge produces 9.5 h operation time
1 h charging gives >3.25 h of operation
Fast charge
3 h from empty to full charge produces 9 h operation time
1 h charging gives 4 h of operation
5.2 Assembling the probe system
WARNING
Risk of explosion due to not approved probe parts.
The original INFICON Hand probe is the only part of the probe system that is certified for use in potentially explosive areas. Other parts of the probe system, or third party accessories are not allowed in potentially explosive areas.
► Pay proper attention to safety when using other accessories for locating and pinpointing gas leaks in none classified areas.
NOTICE
The assembly of the probe system influences the response time stated in the technical data.
The probe system consists of the hand probe, the probe rod, the probe rod extension, the carpet probe and the bell probe. Each probe part is equipped with quick couplings and can be assembled according to the following graphics. The connection to the leak detector is made with the hand probe in all configurations.
The proper functioning of the system can only be assured with compatible probe systems of the manufacturer.
1 Attach the probe rod to the carpet or the bell probe.
38 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Getting started | 5
2 Attach the hand probe to the probe rod.
3 Connect the probe tube of hand probe to the gas inlet of the leak detector, see
"Design of the leak detector [ } 15]".
Install the probe rod extension
To improve the ergonomics of the carpet probe, you can install the "probe rod extension".
1 Remove the male quick fitting (1) from the carpet using as 17 mm wrench.
2 Remove the tape covering the female thread of the extension rod (2).
3 Place the O-ring, found under the tape, flat on the bottom surface of the female threaded hole.
4 Screw the removed male quick fitting (1) in place in the extension rod (2). Make sure the O-ring stays in place by holding the extension rod (2)vertical. If desired you can use nut locking fluid to secure the fitting (1).
IRwin-Operating-instructions-mina66en1-15-(1904) 39 / 128
5 | Getting started INFICON
5 Remove the O-ring and clean the threaded hole on the black fork on the carpet
(where you removed the connector).
6 Place the O-ring flat on the bottom surface of the threaded hole.
7 Screw the extension rod in place in the carpet fork. Make sure the O-ring stays in place by holding the fork vertical.
IMPORTANT: Be careful when tightening. The fork is plastic and the threads are easily destroyed. If desired you can use nut locking fluid to secure the extension rod (2).
8 Connect the longer rod to the carpet.
5.3 Transfer data to computer
NOTICE
System requirements: Windows 7 or later with Bluetooth interface.
40 / 128
5.3.1 Install software for data transfer
1 Install the IRwin App software on your computer.
2 Follow the instructions on the screen.
More complete information about supporting software can be found in separate documents.
5.3.2 Pairing the leak detector with computer via Bluetooth interface
ü Set IRwin in one of the following operating modes: Bar Hole, Gas Purity, House,
Above Ground and Confined Space. Bluetooth is active in all these modes.
ü Make sure IRwin App is installed on your computer.
ü Make sure Bluetooth is activated on your computer.
1 Establish a Bluetooth connection. Refer to the instructions on your computer.
When searching for your detector on the computer, locate the correct serial number of your detector and select that device on the computer. Some computers will prompt you to acknowledge that a connection "password" has been correctly transferred to the detector. You must confirm this on both computer and detector.
2 Start the program IRwin App on your computer.
3 In the IRwin App on your computer, Click Connection (menu item) > Port and select your detector.
IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Getting started | 5
ð The serial number of the leak detector is displayed in the drop down menu.
4 Select "Connect" on the menu bar in IRwin App.
5 Click "Start Streaming" on the menu bar of IRwin App.
5.3.3 Transfer log files from leak detector to computer
ü You have installed IRwin App software on your computer.
ü You have recorded at least one log fil in your detector.
ü You have activated the Bluetooth interface on your computer.
1 Switch on the leak detector
2 Perform self-test.
3 Make sure IRwin is not in Ex or ExTox mode. Bluetooth is active in all other modes.
4 > > Data output > Files (Tab)
5 Start IRwin App on your computer
6 Select the serial number of your detector in “Port” drop down menu. If not found, check setting of Bluetooth interface and repeat paring sequence if necessary.
7 Click “Connect” in IRwin App.
8 Select tab “File” in the graphic window in IRwin App.
9 Select the file to transfer. You can only transfer one file at the time.
10 Click “Download File” in IRwin App.
11 IRwin App will confirm when file is transferred.
12 To delete a file, select the file in the IRwin App and click “Delete File”. Reenter the file list in IRwin to confirm that the file is deleted.
NOTICE
Avoid transferring files while performing a GC Ethane Analysis as this can severely affect the GC analysis.
5.3.4 Real-time streaming from leak detector to computer
ü You have installed IRwin App software on your computer.
ü You have activated the Bluetooth interface on your computer.
1 Switch on the leak detector.
2 Perform self-test.
IRwin-Operating-instructions-mina66en1-15-(1904) 41 / 128
5 | Getting started INFICON
3 Set the leak detector in the desired mode. This must be one of the following operating modes: Bar Hole, Gas Purity, House, Above Ground or Confined
Space. Bluetooth is active in all these modes.
4 Start IRwin App on your computer.
5 Select the serial number of your leak detector in “Port” drop down menu. If not found, check setting of Bluetooth interface and repeat paring sequence if necessary.
6 Click “Connect” in IRwin App.
7 Select tab “Live” in the graphic window in IRwin App.
8 Click “Start Streaming” in IRwin App.
ð The measurement data is now presented as a graph.
While streaming you can change mode as desired.
Streamed data cannot be saved.
Fig. 10: Streamed measurement data
42 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON
6 Operation
WARNING
Danger to life and limb
Due to incorrect and irregular inspections.
► Check all parts of the entire measuring system regularly for damages.
Operation | 6
6.1 Switch ON
WARNING
Explosion hazard
IRwin S and any kind of equipment for charging (for all models), are not ex-protected.
► Do not use the IRwin S or any kind of charger for the S and SXnn versions in explosive atmospheres.
The start-up sequence can differ slightly between the individual models.
1 Press the on / off button on the lower right of the display ("Design of the leak detector").
ð During start-up, the leak detector passes through a warm-up phase of 2 minutes maximum and performs a self-test of software and parameter memories, battery, display, pumps, valve(s), buzzer, sensors and status LED.
2 To confirm the function of the buzzer, press the button next to check mark
(upper right).
3 The status LED lights up in various colors as indicated on the screen. Confirm each color by pressing the check mark (upper right).
4 Confirm the results of the self test.
ð The measurement screen appears.
5 If you want to check the condition of the leak detector even further, perform a calibration test ("Calibration") or a function test ("Performing function tests") before starting to work.
IRwin-Operating-instructions-mina66en1-15-(1904) 43 / 128
6 | Operation INFICON
6.2 Initial setup
IRwin is delivered set in English language, and modes according to German DVGW regulations.
After 2 minutes of inactivity the menu system exits automatically to the measurement screen.
6.2.1 Select available operating modes
IRwin can be also set for general use in "Universal mode" and in several languages.
Enabling Universal mode gives you one mode for the full range of gas concentrations from 1 ppm to 100 volume %.
1 > > General > Custom (Tab)
2 Select Universal mode and press the button next to .
3 Leave the view via .
6.2.2 Adjust the local time
If the time of the leak detector and the local time do not match you can change the time zone.
1 > > General > Time/Date (Tab).
2 Adjust the local time zone.
ð
For the settings, see "Factory settings [ } 32]".
3 Leave the view via .
6.2.3 Setting the language
1 > > General > Misc (Tab) > Language
2 Select a language.
3 Leave the view via .
44 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
6.2.4 Change permission level
The leak detector has three permission levels “Basic“, “Intermediate“ and “Advanced“.
The leak detector starts with the permission level "Basic". The permission levels
“Intermediate“ and “Advanced“ are protected by a password. You can see which
permission level is active in the status line of the display ("Display [ } 19]").
Symbol Permission level
Basic
Intermediate
Advanced (Supervisor)
Password none
1111
1422
Table 5: Permission levels
Change permission level
1 > > Password > Login (Tab).
2 Use the navigation buttons to enter the password for the desired permission level.
3 Leave the view via .
Log out from a higher permission level
If you want to restore the permission for "Basic", you must log off.
1 > > Password > Logout (Tab).
2 Press the check mark button two times to log out.
6.2.5 PPM Adjustment Factor
The PPM Adjustment Factor can be used to slightly overestimate low natural gas concentrations. This can be handy if bump test requirements require an indication of at least 10 ppm.
The selected value (1.0 to 2.0) will be multiplied with the calibrated reading.
Setting this factor to e.g., 1.4 will make IRwin display a 10 ppm reading as 1.4 x 10 =
14 ppm.
This adjustment affects values up to 10 ppm and turns into a minor offset above 10 ppm.
► > > General > Misc (Tab).
6.2.6 Select Gas (Ex)
If desired you can set IRwin to measure Propane (C
3
H
8
) or Butane (C
4
H
10
) instead of methane (CH
4
).
IRwin will use your methane calibration and built-in correction data for the selected gas.
► > > General > Misc (Tab).
IRwin-Operating-instructions-mina66en1-15-(1904) 45 / 128
6 | Operation INFICON
6.2.7 100% CH4 LEL conc
You can select the LEL level according to local standards. Choose between 4.4% and
5.0%.
► > > General > Custom (Tab)
6.2.8 Universal Mode
IRwin can be set up in two general ways: Default and Universal.
The default mode is adapted to German and Dutch work procedures with specific test modes for different applications.
By enabling the Universal mode you can instead combine most of these modes into one Universal mode covering the entire concentration range from ppm to 100 Vol%.
► > > General > Custom (Tab)
6.2.9 AbG Cal Setup
Above Ground Calibration Setup is used to customize the calibration routing for low concentration methane measurements. The Classic setting uses the calibration routine specified by German and Dutch work procedures whereas the Generic option makes it possible to use only one or two gas concentrations instead of three.
Note that reducing the number of gas concentrations may result in lower accuracy not fulfilling the entire specification.
► > > General > Custom (Tab)
Selection of gas concentrations to use is made by changing the Setting of calibration
gas concentration. See "Calibration [ } 49]Setting of calibration gas concentrations
6.2.10 AbG Fnc Test conc
The gas concentration used for Above Ground Function Test can be set to suit your application. You can chose between 10 ppm, 500 ppm and 2.2 or 2.5 Vol%
(representing 50% LEL as set).
► > > General > Custom (Tab)
46 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
6.2.11 Screensaver timeout
Switch on screen
You can activate a screen saver that will turn the screen off if there is no alarm or key action within an adjustable time.
1 > > General > Display (Tab)
2 Select the desired timeout or select "Off" to disable the screensaver.
3 Leave the view via .
► To switch the screen back on, press any button. It will switch on automatically if a gas alarm is triggered.
IRwin-Operating-instructions-mina66en1-15-(1904) 47 / 128
6 | Operation INFICON
6.2.12 Configure data logging
You can record and store your survey with following data in the leak detector.
• Date
• Time
• GPS coordinates
Measurement data of the activated mode:
• CH
4
and CO
2
measurements
• % LEL
• CO, O
2
and H
2
S measurements (IRwin SXT and IRwin SXGT)
The measurement data recording must be configured before the initial use:
ü Login as Intermediate or Advanced
1 > > Data output > Log to File (Tab)
2 Select "Enable log to file" and press the button next to .
3 Select “Log interval” and select a time interval. A blue button is shown in the lower right corner when data logging is enabled. Press the on / off button to start and stop logging to file.
The shorter you set the recording interval the more disk space is required. The percentage of disk space used is shown by the bar below, "Used disk space" on the
Log to File tab.
6.2.13 Enable dead band suppression (IRwin SXnn only)
For the toxic sensors in the application mode “Ex Tox mode“ a dead band suppression can be activated. If you activate the dead band suppression, the leak detector will suppress small signal fluctuations around the zero point of the measurement signal.
1 > > General > Misc (Tab) > dead band Suppression.
ð The symbols (top right) and (bottom right) appear.
2 Press the button next to .
6.2.14 Setting the Function Test Interval
Setting time interval for
Function Test
Reminder
Consider relevant standards and regulations.
ü Login as Intermediate or Advanced
1 > > Function Test > Interval (Tab)
2 Select the function test and the time interval.
ð IRwin S: You can set a time interval.
48 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
ð IRwin SXnn: You can set the time interval individually for the modes “Above
Ground“ / ”Universal (low range)", “Bar Hole” / ”Universal (high range)”, “Ex“,
“Tox Mix“ and “GC”.
3 Leave the view via .
6.2.15 Calibration
The built-in calibration routines guide you through the adjustment of the leak detector sensitivity to the various test gases. You can set calibration intervals for the different adjustments and will then get a reminder when calibration is due ("Adjust calibration reminder").
Allow IRwin to warm up for 1 hour before calibrating for best accuracy.
Recommended calibration intervals
The calibration intervals for the operating modes Ex and Ex Tox are based on the TÜV certificates for these operating modes. For the other modes these time intervals are chosen to give you the best performance. Use the results of the various function tests to judge whether a calibration is needed. If you want to perform more than one calibration procedure at a given time you should run them in the following sequence for best performance.
1. Above Ground or Universal (low range) if Universal mode is enabled
2. Ex CH
4
(IRwin SX, SXT, SXG and SXGT only)
3. Tox (IRwin SXT and SXGT only)
4. Bar Hole CH
4
or Universal (high range) if Universal mode is enabled
5. Bar Hole CO
2
or Universal (20% CO
2
) if Universal mode is enabled
• Above Ground CH
4
/ Universal (low range) - weekly
For calibration in Above Ground mode you need the following gas concentrations:
10 ppm CH
4
, 100 ppm CH
4
, 2.2% CH
4
.
With this calibration you calibrate the higher range (Vol%) of the modes House and Bar Hole at the same time.
• Tox - weekly
For calibration in Ex Tox mode you need Tox mix.
The CO
2
calibration of other modes is not affected. CH
4
of the operation mode Ex is also calibrated during the Ex Tox calibration. However, for best performance, it is recommended to perform the Ex calibration after the Ex Tox calibration.
• Ex CH
4
- weekly
For calibration in Ex mode you need the following gas concentrations:
2.2% CH
4
.
IRwin-Operating-instructions-mina66en1-15-(1904) 49 / 128
6 | Operation INFICON
• Bar Hole CO
2
/ Universal (20.0% CO
2
) - monthly
For calibration of the CO
2
measurement of Bar Hole you need the following gas concentrations: 20% CO
2
.
This calibration does not influence the toxic CO
2 calibration (ppm) in the Ex Tox calibration routine.
• Bar Hole CH
4
/ Universal (high range) - monthly
For calibration of the CH
4
measurement of Bar Hole you need the following gas concentrations:
2.2% CH
4
, 100% CH
4.
Beside the calibrations of the application mode “Bar Hole” the application modes
“Gas Purity” and “Confined Space” will be calibrated in the higher measurement range (Vol%), too.
WARNING
Explosion, asphyxiation, toxic risk
Incorrect calibration leads to incorrect measurement results. Thus you are not properly warned against any possibly dangerous gas concentrations.
► Calibrate the leak detector only when you have been properly trained.
► Calibrate the detector in well ventilated rooms or outdoors.
The methane calibration is possible with gas mixtures in synthetic air only.
The natural background of CO
2
(typically 400 ppm) in ambient air will offset the calibration.
Fresh air exposure should, however, always be by ambient air.
After the leak detector has been calibrated with 100% methane it shows a few ppm increase in the measured value for a few minutes. This is a normal behavior of the leak detector and due to gas residues inside the leak detector.
Retrieving information about the calibration
The leak detector saves the date and the time of the last successful calibration.
► > > Calibration > Info (Tab)
50 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
6.2.15.1 Standard calibration procedures
Prepare calibration
It is not necessary that you calibrate all modes at the same time.
Mode Above
Ground
Calibration Routine
Ex Tox
Universal
(low range)
Bar Hole
CH
4
Universal
(high range)
Bar Hole
CO
2
Universal
(CO
2
)
Above Ground
Universal (low range)
Bar Hole
Universal (high range)
Confined Space
House
Gas Purity
Ex
Tox
X
X
X
X
X X*
X*
X
X
X
*Ex Tox calibration routine also calibrates Ex mode.
No need to run Ex calibration routine separately if Ex Tox calibration has been performed.
X
X
X
Calibrate Above Ground / Universal (low range) first for best performance.
You don’t need to perform any of the calibration routines more than once for a complete instrument calibration.
We strongly recommend using so called ODFRs (On Demand Flow Regulators) to administrate calibration and test gases to the instrument. These regulators automatically supplies the exact gas flow aspired by the instrument and automatically shuts off when disconnected from the sample input.
It is also possible to use other calibration stations for the calibration. However, the following requirements have to be fulfilled.
• The gas flow delivered by the calibration station must be higher than the flow aspired by the leak detector. Set the gas flow to at least 80 l/h.
IRwin-Operating-instructions-mina66en1-15-(1904) 51 / 128
6 | Operation
Calibration
INFICON
• The gas must be delivered at ambient pressure (i.e., do not push gas into the leak detector). The normal way to accomplish this is to create an open connection between gas container and detector inlet so that the excess calibration gas is vented to the ambient. This exhaust should be placed outside the test room, preferably outdoors.
ü Login as Intermediate or Advanced
► > > Calibration > Calibration (Tab)
ð Select the calibration to perform and click the check mark button.
ð The leak detector leads you through the process by indicating which gas to deliver to the detector in each step. Make sure that no residual gases are present in the ambient during Fresh Air step. Most common problems occur if you used 2.2 Vol% or 100 Vol% methane shortly before the trying to calibrate with 10 ppm or 100 ppm methane.
ð A successful calibration is confirmed by a green check mark.
ð A failed calibration is indicated by a red cross.
You can continue to use the leak detector even if calibration failed. The previous calibration parameters are restored.
6.2.15.2 Setting of calibration gas concentrations
Accuracy can be affected by calibration gas tolerance.
For optimum performance adjust the concentrations of the calibration gases set in the leak detector to the actual composition of your calibration gases.
For optimum performance of your leak detector you should use the proposed calibration gases. The exact concentrations of the gas delivered often deviates slightly from these values due to tolerance factors etc. The exact value is typically stated on the analysis certificate of the respective gas. You should therefore adjust the concentrations of the calibration gases set in the detector to match the actual
composition of your calibration gases, see "Concentration of calibration and test gases
ü Login as Intermediate or Advanced
1 > > Calibration > Settings (Tab)
2 Select calibration routine first and then select the gas to adjust (if several).
3 Confirm your selection with .
ð A window appears with controls for adjusting the gas selected.
52 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
4 Set the exact gas concentration value via drop-down menus. The leak detector shows the gas to adjust in the window and the currently stored concentration appears just below the setting line. You cannot set the concentration outside the range.
5 Select "Set" to confirm new value or "Reset" to restore the default value.
The "names" for the different calibration routines is not affected by the adjustment of calibration gas concentrations. The texts displayed during the calibration process will, however, show the adjusted values.
If using the same gas in several calibration routines, you must adjust the concentration of that gas in all relevant routines.
6.2.15.3 Adjust calibration reminder
Activate calibration reminder
You can set an Interval for a calibration reminder for each of the calibration types. The
Interval is set to Off as default, meaning no reminder.
ü Login as Intermediate or Advanced
1 > > Calibration > Interval (Tab)
2 Select the operation mode for which you want to set a calibration reminder and set the desired period of time.
3 Leave the view via .
The leak detetctor saves the set period of time automatically.
IRwin-Operating-instructions-mina66en1-15-(1904) 53 / 128
6 | Operation
6.3 Operating modes
After 2 minutes of inactivity the menu system exits automatically to the measurement screen.
INFICON
6.3.1 Select operating mode
After finishing the self test the leak detector goes into the operating mode.
► Press repeatedly to scroll through the operating modes.
The detector is ready to use when the measurement value(s) turn(s) black and the status LED illuminates blue (green for Ex and Ex Tox modes).
6.3.2 Operating mode “Above Ground”
This mode is the most sensitive and intended for pipeline survey. This is done by taking samples on the ground surface along the pipe route typically using the carpet probe.
The detector is ready to use when the measurement value(s) turn(s) black and the status LED illuminates blue (green for Ex and Ex Tox modes).
WARNING
Do not use Above Ground for safety check.
The accuracy is not certified for LEL measurement.
Observe all safety precautions when alarms are given.
Never open or enter a confined space without first testing LEL level in the Ex or ExTox mode.
54 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON
1
Operation | 6
1
6
3
1
2
5
9
7
(10 PPM)
Measurement range
7
9
1
Status indicator LED
Buzzer
2
8
5
Above ground
Fig. 11: “Above Ground” display
1
3
5
Navigation buttons
Battery status indicator
Measured value, target gas and unit
2
4
6
Navigation and on / off button
Mode icon
Analog bar indicating range (grey), measurement value (blue) and alarm settings (red lines) in quasilogarithmic scale
8 Operation mode and menu bar
10 PPM threshold in Above Ground and Confined Space modes.
Measurement range 1 ppm - 5 Vol% methane
IRwin-Operating-instructions-mina66en1-15-(1904) 55 / 128
6 | Operation INFICON
6.3.3 Operating mode "Universal"
This mode combines highest sensitivity for PPM levels of Methane all the way up to
100 Vol% and can be used for all applications except safety checks.
Use Ex or ExTox mode for checking the safety of the work place.
WARNING
Do not use Universal mode for safety check.
The accuracy is not certified for LEL measurement.
Observe all safety precautions when alarms are given.
Never open or enter a confined space without first testing LEL level in the Ex or ExTox mode.
WARNING
Gas alarms are not safe in this mode!
The background concentration around the detector is zeroed out. Signal will only be displayed if probe detects higher gas than the reference inlet.
The accuracy of the methane measurement in the operating mode "Universal" is not certified for the LEL measurement.
► Observe all safety precautions when alarms are given.
56 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
1
1
2
3
CH
4
ppm
4
(10 ppm)
5
Universal
Measurement range
Fig. 12: Universal
1
3
5
Measurement value
Unit of measured value
Zeroing button, if needed. Zeroing symbol disappears if gas is detected.
2
4
6
Measurement range
Measured gas
PPM threshold
Data logging activation.
1 ppm - 100 Vol% methane
6
IRwin-Operating-instructions-mina66en1-15-(1904) 57 / 128
6 | Operation INFICON
6.3.4 Operating mode “Bar Hole”
This mode is intended for pin-pointing leaks by measuring the concentration in the ground. The detector is ready to use when the measurement value is black and status
LED is blue. No gas alarms are given in this mode.
WARNING
No gas alarms are given in the Bar Hole mode.
When pin-pointing leaks you will often encounter very high concentrations, far above the LEL limit. Be careful and follow all relevant safety regulations.
58 / 128
1 4
3
2
5
6
7
8
Bar hole
3
5
7
Fig. 13: "Bar hole" display
1 Analog bar showing range (grey) and measurement value (blue) in
Vol% (quasi logarithmic scale)
Measured gas
Unit of the measured value
C
2
H
6
compensation value
2
4
6
8
Measurement value
Symbol of active operating mode
CO
2
measurement value
Start Automatic Bar hole test.
Zeroing button, if needed. Zeroing symbol disappears if gas is detected.
IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON
Measurement range Measurement range
Operation | 6
0.5 Vol% - 100 Vol% CH
4
0.1 Vol% - 20 Vol% CO
2
You can readjust the zero point of the CH
4
Vol% value if needed. This can be done if deviation is within ± 5 Vol%. Press lower left button if "0" appears in blue.
Deviation outside ± 5 Vol% is an indication that the system needs to be calibrated.
NOTICE
Measurement errors due to incorrect ethane compensation
Incorrect ethane compensation value can result in significant deviations in the estimated methane concentration.
See Entering ethane concentration manually [ } 76] for information on how you set
the value for ethane compensation. See also Setting reminder for IR Ethane Analysis
6.3.4.1 Automatic “Bar Hole” testing
The detector has an automated function for ventilating and testing the gas content in bar holes. This function will take three samples in the hole with a pre-determined evacuation time between the samples. Default setting is 10 s measurement and 3 minutes evacuation between samples. You can also set the detector to terminate the sequence when the CO
2
level has dropped below a given threshold even if all three samples are not yet taken (Setting parameters for “Auto Bar Hole”).
A low content of CO
2
is a good marker for a well ventilated hole.
The leak detector starts beeping when the test sequence is finished and displays the methane and CO
2
concentrations.
1. Set the leak detector in Bar Hole mode.
2. Press the lower left key „ “ to start Auto Bar Hole test.
3. Place the probe in or on the hole an press the check mark key.
4. Leave the leak detector to complete the Auto bar Hole test.
5. You can read the result of the individual measurements on the display when the sequence is ready. The leak detector will start beeping when the test is ready.
Note that three results will be presented even if IRwin did not make all three. The results are cleared when you press the check mark key.
IRwin-Operating-instructions-mina66en1-15-(1904) 59 / 128
6 | Operation INFICON
6.3.4.2 Setting parameters for “Auto Bar Hole”
1 > > Auto Bar Hole > Settings (Tab)
2 Measuring Duration. This is the time the detector tests the gas in each measurement. Select 10 - 20 s (5 s increments)
3 Evacuation Duration. This is the pumping/evacuation time between measurements. Select 3, 4, 5, 10 or 15 min.
4 CO
2
Limit on Evacuation. This is the limit indicating a well evacuated test point.
Auto Bar Hole sequence will be terminated directly efter completing one measurement in which CO
2
level has dropped below the set limit. Select Off to disable function or 1 - 5% to activate function.
60 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
6.3.5 Operating mode "GC Ethane Analysis"
This mode is intended for determining if a gas sample is natural gas (NG) or liquefied petroleum gas (LPG), by identifying the methane, ethane and propane content.
Natural gas (NG) typically contain at least 0.5 vol% of ethane in addition to a high concentration of methane.
Liquefied petroleum gas (LPG) typically contain propane or butane or a mixture of the two.
These differences are the basis for identifying gas type from an hydrocarbon analysis.
While significant ethane content (>0.5 Vol%) in a methane rich mixture is a certain indication of natural gas, it is not entirely safe to say that no indication of ethane in a methane rich mixture is a certain indication of swamp gas.
Some natural gas contains no or only traces of ethane.
This can be due to several reasons, for example:
• The gas source is solely or partly liquefied natural gas (LNG) which sometime has very low ethane content.
• The gas source is solely or partly biogas.
• The natural gas source has very low ethane content.
1
2
3
4
6
5 GC Ethane Analysis
Fig. 14: GC Ethane Analysis
IRwin-Operating-instructions-mina66en1-15-(1904) 61 / 128
6 | Operation INFICON
Measurement range
1 Analog bar indicating concentration range (grey) needed to perform GC Ethane Analysis and measurement value (blue) in quasi logarithmic scale
Mode icon
2
4
Battery status indicator
3
5 Operation mode and menu bar 6
Measured value, target gas and unit
Take sample for analysis.
Becomes active (blue) and cross disappears when natural gas concentration is high enough.
Detectable range ≥ 0.5 Vol% ethane in gas.
The following procedure is highly recommended on site to strengthen the evidence that a gas sample is indeed natural gas:
• Check that ambient temperature is within range for GC Ethane Analysis: 0 – 50°C
(32-122°F)
• Perform a GC Ethane Analysis Function test to check that the GC Ethane Analysis can detect ethane down to 0.5 Vol% in the gas sample. Gas used see
Concentration of calibration and test gases [ } 34]
• Make a GC Ethane Analysis on a sample of natural gas from the actual gas line at your site.
• Make a GC Ethane Analysis on the gas to be identified and compare the resulting curve to that obtained from the natural gas sample. Three consecutive tests with same result should serve as a clear result.
• If any doubt on this result remains perform a final function test to assure that the unit was working properly during testing.
If the ethane content in the natural gas is below 0.5 Vol% the GC Ethane Analysis cannot be used. The GC must also give a clear indication of ethane to serve as an indicator of natural gas. A clear indication is a ticked ethane box on the screen.
The IRwin GC is sensitive enough to detect 0.5 % ethane in methane.
The operator is fully responsible for evaluating the logic and likelihood of the correctness of the result being presented by IRwin, based on the ethane content in the tested gas, the maintenance and functionality of the IRwin Methane Leak Detector and other circumstances related to the issue.
62 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
NOTICE
Overloading the GC column with high gas concentrations will make analysis slow and possible non-functional
Move instrument away from leak area to fresh air directly after taking sample to avoid this. If you suspect overloading, you can make a prolonged column purge by pressing the sampling button (lower left) more than 3s.
Purging will continue until manually terminated or until instrument shuts down.
1 When in GC Ethane Analysis mode, expose to gas
2 Press lower left button when cross disappears and sampler symbol turns blue.
3 Confirm that you want to perform GC Ethane Analysis and follow instructions:
ð The detector prompts you to keep exposing to gas
4 Move probe to fresh air, when the detector prompts you to place probe in fresh air.
ð The semiconductor sensor is purged to prepare for GC Ethane Analysis.
ð GC analysis starts and the resulting graph is being displayed. Remaining time is displayed.
5 Press lower left button (x) and confirm to end the analysis when ethane analysis is completed (typically 80s). Propane is not analyzed if ending.
ð Result is interpreted and displayed (see A and D below)
ð The GC system is purged. Remaining purge time is displayed (see B below).
6 Press lower right to zoom graph (see C below)
7 Press lower left button to save analysis (see E below).
8 You can postpone purging of GC system by changing mode. Confirm by pressing Check mark button.
ð Purging will restart automatically if shutting IRwin down or attempting to make a new GC Ethane Analysis.
IRwin-Operating-instructions-mina66en1-15-(1904) 63 / 128
6 | Operation INFICON
A
B
D
NG with Ethane detected
E
GC Ethane Analysis
Fig. 15: NG with Ethane detected
C
The data saved after a GC Ethane Analysis contains data for when and where the sample was taken.
The data can be retrieved by the IRwin APP (Windows application) that will create a
PDF report for the analysis.
You can save a large number of analysis data files for later report generation.
The GC Ethane Analysis requires no calibration.
NOTICE
Avoid transferring files while a GC analysis is running as this can severely affect the result
64 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
6.3.6 Operating mode “Confined Space”
WARNING
The accuracy of the methane measurement in the operating modes "Confined
Space" and "House" is not certified for the LEL measurement.
Observe all safety precautions when alarms are given. Never open or enter a room without first checking the LEL level in Ex or Ex Tox operation mode.
The operating mode House is not intended for pre-entry safety checks. The fire potential can only be determined in one of the Ex or Ex Tox operating modes.
NOTICE
The reference inlet of the leak detector must be in fresh air.
Only probe inlet should be in the confined space.
This mode is used for checking the gas concentration in confined spaces such as man holes, duct openings, etc. The leak detector is ready to use when measurement value is black and the status LED is blue.
This mode is not available if Universal mode is activated.
IRwin-Operating-instructions-mina66en1-15-(1904) 65 / 128
6 | Operation INFICON
1
2
3
4
5
Measurement range
Confined Space
Fig. 16: “Confined Space” display
1
3
5
Analog bar indicating range (grey), measurement value (blue) and alarm settings (red lines) in VOL%
(quasi-logarithmic scale)
Measured gas
Unit of the measured value
2
4
Measurement value
Symbol of active operating mode
Measurement range 1 ppm - 100 Vol% methane
66 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
6.3.7 Operating mode "House"
WARNING
Gas alarms are not safe in this mode!
The background concentration around the detector is zeroed out. Signal will only be displayed if probe detects higher gas than the reference inlet.
The accuracy of the methane measurement in the operating modes "Confined Space" and "House" is not certified for the LEL measurement.
► Observe all safety precautions when alarms are given.
This operating mode is used to check installations and pinpoint leaks in buildings. The detector is ready to use when the measurement value is black and status LED is blue.
This mode is not available if Universal mode is activated.
1
2 3
4
5 6
Measurement range
House
Fig. 17: “House” display
1
3
5
Analog bar indicating range (grey), measurement value (blue) and alarm settings (red lines) in VOL%
(quasi-logarithmic scale)
Measured gas
Unit of the measured value
2
4
6
Measurement range
Measurement value
Symbol of active operating mode
Set point for PPM alarm
1 ppm - 5 Vol% methane
IRwin-Operating-instructions-mina66en1-15-(1904) 67 / 128
6 | Operation INFICON
6.3.8 Operating mode "Gas Purity”
This mode is intended for checking that the gas has reached a certain point in the network when a previously purged section is refilled.
This mode is not available if Universal mode is activated.
WARNING
No gas alarms are given in this operation mode!
This operating mode is intended for the measurement of higher gas concentrations. In this mode you usually see up to 100 Vol% methane.
Follow all relevant safety precautions and regulations.
1
2 3 4
5
6
Measurement range
3
5
7
1
0 7
Fig. 18: “Gas Purity” display
Gas Purity
Analog bar indicating range (grey) and measurement value (blue) in
Vol% (quasi logarithmic scale)
Measured gas
Unit of the measured value
Zeroing button, if needed. Zeroing symbol disappears if gas is detected.
2
4
6
Measurement value
Symbol of active operating mode
C
2
H
6
compensation value
Measurement range 0.1 - 100 Vol% methane
68 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
You can readjust the zero point of the CH
4
Vol% value if needed. This can be done if deviation is within ± 5 Vol%. Press lower left button if "0" appears in blue.
Deviation outside ± 5 Vol% is an indication that the system needs to be calibrated.
IRwin-Operating-instructions-mina66en1-15-(1904) 69 / 128
6 | Operation INFICON
6.3.9 Operating mode "Ex"
This operating mode is used to check or monitor the fire or explosion risk in the working environment. The concentration of the gas is given in % LEL. A signal of
100% LEL or higher (> % LEL) means there is a risk of fire.
This mode is not available if Universal mode is activated.
WARNING
Explosion risk
Be aware that the displayed concentration represents the concentration at the gas inlet point of the probe. The concentration can be much higher at another point close by.
The leak detector is ready to use when the measurement value is black and the status
LED is green and the acoustic alive signal beeps every 20 seconds. See warning text.
1
2
3 4
70 / 128
5
0 6
Ex
Fig. 19: “Ex” display
1
3
Analog bar indicating range (grey), measurement value (blue) and alarm settings (red lines) in Vol%
(quasi-logarithmic scale)
Measured gas
2
4
Measurement value
Symbol of active operating mode
IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON
5 Unit of the measured value
Operation | 6
6 Zeroing button, if needed. Zeroing symbol disappears if gas is detected.
±Measurement range
Measurement range 1 - 100 %LEL methane
You can readjust the zero point of the %LEL value if needed. This can be done if deviation is within ± 5 %LEL. Press lower left button if "0" appears in blue.
Deviation outside ± 5 %LEL is an indication that the system needs to be calibrated.
If required, change the primary target gas from methane (CH
4
) to propane (C
3
H
8
) or butane (C
4
H
10
). The leak detector will then recalculate the signals to propane or butane values using a certified correlation curve.
Use this feature if you need to work on LPG installations.
For changing gas type see "Change gas type for %LEL measurement [ } 75]".
Status LED
Keep an eye on the status LED when working in either of the safety certified modes
Ex or Ex Tox. The LED is green when the leak detector works properly and no gas alarm is activated.
Listen also for the acoustic “alive signal” that beeps every 20 s to indicate normal operation if there are no gas alarms.
The status LED changes to red when a gas alarm is triggered or an error exists.
Status LED is yellow in special states such as when in setting menus or calibration etc. Blue status LED means normal operation in all other modes.
WARNING
Explosion risk may exist when "alive" signal not present.
When operating in any of the two safety related modes Ex and Ex Tox respectively, the leak detector will confirm its proper function with a short acoustic "alive" signal every 20 s and by the status LED being green.
► Leave hazardous areas immediately if the alive signal does not sound every 20 s.
► Do not use the leak detector until it operates normally again.
IRwin-Operating-instructions-mina66en1-15-(1904) 71 / 128
6 | Operation
Ex
INFICON
The acoustic signal and the status LED are the primary alarm signals. All gas alarms have a latching function. After a gas alarm is triggered, a check mark key appears in the upper right corner of the display.
You can mute the acoustic alarm by pressing the check mark key. The alarm will come on again if a new alarm is triggered.
After the leak detector has been exposed to high methane concentrations, it often displays increased measurement values in the ppm range for a few minutes. This is a normal behavior of the leak detector due to gas residues inside the leak detector.
72 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
6.3.10 Operating mode "Ex Tox”
This operation mode is used to monitor and warn for explosive and toxic gas concentrations as well as oxygen deficiency in the working area.
Measurement range
1
4
2 3
0 CH
4
%LEL
21.6% O
2
5
0.05% CO
2
6
0.0 ppm H S
2
0 ppm CO
0
Ex Tox
Fig. 20: “Ex Tox” display
1
3
5
Analog bar indicating range (grey), measurement value (blue) and alarm settings (red lines) of flammable gas in Vol% (quasilogarithmic scale)
Measured gas
Unit of the measured value
2
4
6
Measurement value
Symbol of active mode
Measurement values of toxic gases and oxygen in ppm and Vol
% respectively
7 Zeroing button, if needed. Zeroing symbol disappears if gas is detected.
The leak detector shows the gas concentrations of CH
4
, CO
2
, CO, H
2
S and O
2
.
Gas
CH
4
, C
3
H
8
or C
4
H
10
CO
2
CO
H
2
S
Measurement range
1 - 100% LEL
0.1 - 5 Vol%
1 - 500 ppm
1 - 400 ppm
IRwin-Operating-instructions-mina66en1-15-(1904) 73 / 128
6 | Operation INFICON
O
2
0.1 - 25 Vol%
You can readjust the zero point of the %LEL value if needed. This can be done if deviation is within ± 5 %LEL. Press lower left button if "0" appears in blue.
Deviation outside ± 5 %LEL is an indication that the system needs to be calibrated.
If desired you can change the primary target gas from methane (CH
4
) to propane
(C
3
H
8
) or butane (C
4
H
10
). The leak detector will then recalculate the signals to propane or butane values using certified correlation curves.
Use this feature if you need to work on propane or butane installations.
For changing gas type see "Change gas type for %LEL measurement [ } 75]".
Status LED
Keep an eye on the status LED when working in either of the safety certified modes
Ex or Ex Tox. The LED is green when the leak detector works properly and no gas alarm is activated.
Listen also for the acoustic “alive signal” that beeps every 20 s to indicate normal operation if there are no gas alarms.
The status LED changes to red when a gas alarm is triggered or an error exists.
Status LED is yellow in special states such as when in setting menus or calibration etc. Blue status LED means normal operation in all other modes.
Ex Tox
WARNING
Explosion risk may exist when "alive" signal not present.
When operating in any of the two safety related modes Ex and Ex Tox respectively, the leak detector will confirm its proper function with a short acoustic "alive" signal every 20 s and by the status LED being green.
► Leave hazardous areas immediately if the alive signal does not sound every 20 s.
► Do not use the leak detector until it operates normally again.
The acoustic signal and the status LED are the primary alarm signals. All gas alarms have a latching function. After a gas alarm is triggered, a check mark key appears in the upper right corner of the display.
You can mute the acoustic alarm by pressing the check mark key. The alarm will come on again if a new alarm is triggered.
74 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
After the leak detector has been exposed to high methane concentrations, it often displays increased measurement values in the ppm range for a few minutes. This is a normal behavior of the leak detector due to gas residues inside the leak detector.
NOTICE
Measuring of high gas concentrations of H
2
S or CO reduces sensor life time.
Frequent measuring of high gas concentrations will reduce the life time of the sensors for the detection of toxic gases.
Set Time Weighted
Average (TWA)
The leak detector has a TWA-Function (Time Weighted Average Function) with which alarms can be given according to LTEL (Long Term Exposure Limit) or STEL (Short
Term Exposure Limit). The LTEL value and the STEL value correspond to the maximum dose of a substance to which a person should be exposed over a defined period.
• STEL is the weighted average value over 15 minutes (for the respective toxic gas).
• LTEL is the weighted average value over 8 hours (for the respective toxic gas).
To set the Alarm Level for LTEL and STEL, see "Set limits and alarm levels [ } 79]".
NOTICE
The TWA calculations are reset when you switch from Ex Tox to another operating mode.
For changing gas type see "Change gas type for %LEL measurement [ } 75]".
6.3.10.1 Change gas type for %LEL measurement
ü Login as Intermediate or Advanced
► > > General > Misc (Tab) > Select gas
ð Choose between methane (CH
4
), propane (C
3
H
8
) or butane (C
4
H
10
).
This option is only valid for the LEL measurement (operating modes Ex and Ex Tox).
IRwin-Operating-instructions-mina66en1-15-(1904) 75 / 128
6 | Operation INFICON
6.3.11 IR Ethane Analysis and compensation
Incorrect Ethane compensation value can result in significant deviation in the estimated methane concentration, in the operating modes "Above Ground", "Bar
Hole", "Confined Space" and "Gas Purity".
Why an IR Ethane
Analysis?
INFICON single-cell wide range technology
(patent-pending)
This function is mainly used to distinguish between natural gas and swamp gas from underground biological decay. Natural gas contains varying concentrations of ethane while swamp gas does not.
Such analysis is traditionally made with a gas chromatograph. With IRwin the same analysis with similar sensitivity is performed much faster. Instead of the typical 5 to 10 minutes a gas chromatograph requires for an analysis IRwin only needs a few seconds.
The second purpose of the IR-Ethananalyse is to establish the ethane concentration for compensation at high methane levels.
The IR-Ethananalyse can be made from below 1 to 8 Vol% ethane in the natural gas.
The resolution is 0.1% and the accuracy better than 1 Vol%.
IRwin methane leak detector employs the unique INFICON single-cell wide range technology to detect all concentrations from 1 ppm to 100 Vol% methane. This technology is very sensitive to the ethane content in the natural gas. Therefore, large deviations can occur at higher gas concentrations if the ethane concentration is not compensated for.
The leak detector has a built-in routine for ethane compensation. The ethane content can be entered manually, if known. A more practical way is to use the IR-
Ethananalyse function to measure the ethane content and automatically set the compensation concentration. The safety certified modes Ex and Ex Tox are not affected by this phenomenon.
IRwin SXG and SXGT models also have GC Ethane Analysis capability for distinguishing between Natural gas (NG), liquid petroleum gas (LPG) and swamp gas by determining the presence of methane, ethane and propane in the sampled gas.
The GC Ethane Analysis function is not affecting the ethane compensation.
76 / 128
6.3.11.1 Entering ethane concentration manually
If you know the current ethane concentration in your natural gas you can enter this value manually. We recommend, however, that you use the built-in IR-Ethananalyse to determine the ethane concentration as this typically varies from day to day and from location to location.
ü Login as Intermediate or Advanced
IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
1 > > IR Ethane Analysis > Settings (Tab)
2 Select "Ethane (%) for Compensation".
3 Set the desired ethane concentration.
ð The value shown is the current setting. The last value (indicated by an “A”) is the result of the last IR Ethane Analysis.
4 Leave the view via .
6.3.11.2 Setting reminder for IR Ethane Analysis
You can set the detector to remind you to make an IR-Ethananalyse. If set, the detector will ask if you want to perform a new IR-Ethananalyse or keep the current value.
1 > > IR Ethane Analysis > Settings (Tab)
2 Select ”Reminder for Ethane Compensation Analysis”.
3 Select between the following settings:
Off
Start-up
Always
No reminder
Reminder first time you enter "Bar Hole" mode after start-up of detector (10 s delay)
Reminder every time you enter "Bar
Hole" mode (10 s delay).
6.3.11.3 Performing IR Ethane Analysis after automatic reminder
You need a methane concentration in the range of 5 to 15 Vol% to perform the IR-
Ethananalyse. The compensation is only needed for high gas concentrations and is therefore only needed when you have found a significant leak.
You should connect the Bell probe before performing the IR-Ethananalyse.
1 Change to operating mode "Bar hole".
ð If reminder is activated a window will appear showing the current ethane concentration setting. The leak detector asks if you want to keep the current value or a make a new analysis.
2 Select "New” and press the bell firmly down onto the place where you found the gas.
ð A number between 0 and 100 appears on the display.
3 If the value is smaller than 20, push the probe harder onto the surface or look for a place with higher methane concentration. If the value is higher that 80 you should move the bell away from the leak to find a spot with lower concentration.
IRwin-Operating-instructions-mina66en1-15-(1904) 77 / 128
6 | Operation INFICON
ð Try to make the signal stabilize at a value between 20 and 80. When succeeding, the detector will need another 2-3 seconds to calculate the ethane concentration. The result is displayed as % C
2
H
6
.
ð Press check mark to continue. IRwin will ask if you want to "Save Ethane
Concentration for Compensation".
ð Select "No" if analysis was made for confirming natural gas or "Yes" to use for compensation. If selecting "Yes" the value is saved as ethane concentration and shown in the C
2
H
6
field.
See Operating mode “Bar Hole” [ } 58].
ð An ethane concentration value of 1 % or higher is a safe indication of natural gas. A value between 0.4 and 1 % is a probable indication of natural gas.
If the measured ethane values are between 0 and 2% it is recommended that you repeat the measurement at least 3 times on the same position to check the result is consistent.
6.3.11.4 Starting IR Ethane Analysis manually
1 > > IR Ethane Analysis > Start (Tab)
2 Press button next to check mark to start.
3 Press the bell firmly down onto the place where you found the gas.
ð A number between 0 and 100 appears on the display.
4 If the value is smaller than 20, push the probe harder onto the surface or look for a place with higher methane concentration. If the value is higher that 80 you should move the bell away from the leak to find a spot with lower concentration.
ð Try to make the signal stabilize at a value between 20 and 80. When succeeding, the detector will need another 2-3 seconds to calculate the ethane concentration. The result is displayed as % C
2
H
6
.
ð Press check mark to continue. IRwin will ask if you want to "Save Ethane
Concentration for Compensation".
ð Select "No" if analysis was made for confirming natural gas or "Yes" to use for compensation. If selecting "Yes" the value is saved as ethane concentration and shown in the C
2
H
6
field.
See Operating mode “Bar Hole” [ } 58].
ð An ethane concentration value of 1 Vol% or higher is a safe indication of natural gas. A value between 0.4 and 1 Vol% is a probable indication of natural gas.
If the measured ethane values are between 0 and 2% it is recommended that you repeat the measurement at least 3 times on the same position to check the result is consistent.
78 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
6.3.11.5 Viewing IR Ethane Analysis info
Display the date of the last IR-Ethananalyse as follows.
► > > IR Ethane Analysis > Info (Tab)
ð The date and the time of the last C
2
H
6
Analysis will be shown.
ð IRwin SXG and SXGT will also show the time of the last GC Ethane Analysis.
6.3.12 Set limits and alarm levels
WARNING
Explosion hazard
Irwin S is not certified for use in explosive atmospheres and is also not suitable for estimating fire potential (LEL level).
Alarm indications:
Gas alarms are usually separated into three levels of severity:
Priority Alarm
Low AL1
Medium AL2
High AL3
Display and status LED
Red, flashing (1 Hz)
Red, flashing (1.5 Hz)
Red, steady on
Acoustic signal
1 Hz
1.5 Hz
Steady on
The detector will beep and the status LED and the display will be red if the measured gas concentration exceeds a set alarm level (below threshold for O
2
alarms AL1 and
AL2).
Faster flashing or beeping will indicate more severe conditions. See table above.
The oxygen alarm differs from the other in that the main risk (asphyxiation) is associated with low oxygen. The alarms AL1 and AL2 for oxygen warns when the oxygen concentration goes below the respective alarm level. AL3 for oxygen is activated when oxygen concentration exceeds AL3.
The % unit in the settings menu means % of LEL for Ex and Ex Tox operation modes and volume % for all other modes. In the operating modes "Above Ground", "Confined
Space" and "House" you can also set a PPM level alarm to help when surveying for leaks.
The acoustic signal and the status LED are the primary alarm signals of the operating modes "Ex" and "Ex Tox".
IRwin-Operating-instructions-mina66en1-15-(1904) 79 / 128
6 | Operation
Setting alarm thresholds
INFICON
• All alarms are latching and must be confirmed manually even if the alarm conditions that lead to the alarm are no longer existing. Alarms are confirmed by pressing the upper right key, next to the check mark.
The alarm will come on again if the alarm condition is still valid.
• You can temporarily mute the alarm by pressing the lower left key ( ). The muting will turn off automatically after 15 minutes.
You can unmute the alarm manually by pressing the lower left key again.
• Muting will also turn off if a higher alarm is triggered.
ü Login as Intermediate or Advanced
1 > > Thresholds
2 Select desired tab:
ð CxHy (gas survey and LEL alarms)
ð Safety (toxic and oxygen alarms)
ð xTEL (toxic exposure limits)
3 Set the desired thresholds.
4 Leave the view via .
When you press the lower left key you mute the acoustic alarm . You can unmute the alarm manually by pressing the same button again. The alarm will be unmuted automatically after 15 minutes.
6.3.13 Set the brightness of the display
1 > > General > Display (Tab) > Brightness
2 Select a value for the brightness.
3 Leave the view via .
6.3.14 Performing function tests
Correct readings only if settings are correct.
If the leak detector does not work correctly, it may not show every leak.
► Test the leak detector regularly as recommended.
Use the Function tests to test the status and accuracy of the leak detector. The leak detector will guide you through the test procedures and tell you when operator actions are needed.
80 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
Recommended
Function test intervals
(based on IRwin model)
Once you have set the function test intervals, the leak detector will remind you to
perform the respective Function test, see "Setting the Function Test Interval [ } 48]".
• Above ground: Before starting work (10 ppm CH
4
)
• Ex: Daily (2.2% CH
4
)
• Ex Tox: Daily (2.2% CH
4
, CO
2
, CO, O
2
, H
2
S)
You can use different calibration stations for the Function tests. However, the following requirements have to be fulfilled.
• Best choice is to use On Demand Flow Regulators (ODFR) that opens automatically when IRwin pulls gas delivers the exact amount required. If using a conventional pressure regulator type, make sure gas flow through the calibration station is higher than the gas flow of the leak detector.
• The gas must not be fed with over pressure into the leak detector. This is best taken care of by using "ODFR" regulators.
ü To perform the Function test, you need fresh air and either 10 ppm CH
4
, 2.2 Vol%
CH
4
in synthetic air. For IRwin SXT and SXGT you will also need the ToxMix gas mixture.
1 Check the leak detector for visible damage.
2 > > Function Test > Test (Tab)
3 Select the desired function test from the list.
4 Start the selected Function test by pressing .
ð The leak detector guides you through the process.
You can display the date of the last function test.
► > > Function Test > Info (Tab)
This time is the basis for the next function test reminder. Refer to "Setting the Function
Function tests expects factory default concentrations. Using other concentrations, even if within the adjustable calibration concentration range, can make the automatic function test interpretation invalid.
If you fail function test and you are using non-standard gases we recommend that you test the instrument by simply exposing the respective mode for the test gas and observe the reading. This way you can still verify the function but the last passed function test will not be logged if automatic interpretation failed.
6.4 Measuring
ü
Connect the leak detector to the desired probe. See "Function [ } 14]".
1 Switch on the leak detector by pressing the I/O key.
IRwin-Operating-instructions-mina66en1-15-(1904) 81 / 128
6 | Operation INFICON
2 Complete the self test.
3
Make sure that the desired basic settings are correct. See "Initial setup [ } 44]".
4 Make sure that the settings for the desired operating mode are correct. See
5 Switch to the desired operating mode.
6 Consider all relevant regulations for the operation in question.
7
Record the measured values if desired. See "Logging data to file [ } 82]".
8 Start measuring.
6.5 Perform self test
When you turn on the leak detector it will automatically perform a self test for the main components such as software and parameter memories, battery, gas sensors, valve(s), pumps, buzzer and status LED.
If not all test are passed successfully it may be that some operating modes cannot work properly. That is indicated by the status LED being yellow. The test can, for example, fail because you accidentally blocked an inlet during the test.
If the problem has been solved it is possible to re-start the self test sequence and get the detector into normal operation mode without having to re-start the complete system.
► > > Function Test > Test (Tab)
To understand more about the causes of test failure and the leak detector staying in a
special state see “Warnings and error messages [ } 84]".
6.6 Viewing Status menu
You are able to read the current status of the self diagnostic functions of the system in a menu.
When troubleshooting or contacting the service, use this information.
► > > Status
6.7 Logging data to file
Starting/stopping the data logging.
IRwin can log data in all operating modes. This is especially practical to save data from leak surveys in Above Ground mode.
ü
You have configured the data logging, see "Configure data logging [ } 48]".
ü You have set the leak detector in operating mode Above Ground.
1 To start or stop the logging press or in the measurement screen.
2 Confirm your choice.
82 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
6.8 Viewing logfile list
ü Login as Intermediate or Advanced
► > > Data output > Files (Tab)
ð The recorded files are shown in the Log files table.
6.9 Information
During operation, the display shows information that helps you operate the leak detector. Activated operation mode, warnings and error messages are shown in addition to the measurement values.
The leak detector is equipped with extensive self-diagnostic functions. If the electronics detect an incorrect state, the leak detetcor displays this as far as possible and interrupts operation when necessary.
Call up detector information
6.9.1 Retrieve information and statistics
1 > > ABOUT.
2 Leave the view via .
Call up statistics 1 > > STATISTICS
2 Leave the view via .
6.9.2 Viewing the Service Screen
On the service screen you will find details about sensor signals and monitored supply voltages, for example. During contact with the service department this could be useful.
ü Login as Advanced
► > > Service Screen
6.10 Activating Auto-Rotate Screen
When you set the auto-rotate screen the display will flip automatically as soon as you turn the leak detetctor upside down.
IRwin-Operating-instructions-mina66en1-15-(1904) 83 / 128
6 | Operation INFICON
1 > > General > Display (Tab) > Auto-rotate screen
ð The symbols (top right) and (bottom right) appear.
2 Select auto-rotate screen press button next to crossed box .
ð To switch off the auto-rotate screen, press button next to empty box .
Switching OFF
Hard switch off
6.11 Switching OFF
1 Press the "On / Off” key for 3 seconds.
ð The leak detector asks whether you want to switch off the instrument.
2 Select “Yes” and press the check mark key.
Some errors can make the leak detector go into watchdog state. The acoustic alarm beeps continuously and the status LED is red. The leak detector may not even be reacting to the menu buttons. If this happens you must turn the leak detector off. Press and hold the "On / Off” key until the detector shuts down (roughly 9 s).
Do not use this option without cause. You risk losing settings and the bias of the oxygen sensor on SXT and SXGT models. This leads to a prolonged warm up at next power on. It can take up to 12 hours to stabilize the oxygen sensor if sensor bias is lost form more than a few minutes. If you need to do a hardware switch off you should restart the leak detector immediately, If the leak detector malfunctions again you will need to contact INFICON service.
1 Press the "On / Off” key for 9 seconds.
2 Restart the leak detector as soon as possible to restore O
2 sensor bias (IRwin
SXT and SXGT).
Warnings
6.12 Warnings and error messages
Warnings are given for conditions that can impair the accuracy of measurements. The operation of the leak detector is normally not interrupted but some functions may be barred or indicated as non-compliant by status LED being yellow and values being red instead of black.
If the navigation button “Menu“ is colored red this means that there has been an error that you have not yet checked. If the cause of the error remains there will be a new warning/error alarm and a new entry in the error log. The navigation button turns blue again when you have read the warnings list.
Open the warnings log list:
► > > Warnings
Error messages Errors are events that force the interruption of the operation. Errors are also logged in the warnings log list.
84 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
The division between errors and warnings is sometimes not very clear. In both cases it is worth restarting the detector to try to pass the self -test again. Charging the battery also reduces the risk for some errors and warnings.
No.
Message
Warnings
Red LED / Constant beep
Possible error sources Remedy
HW Watchdog triggered Perform hard switch off.
Restart detector.
W2
W3
W4
W5
W6
W7
W8
W9
W10
W11
W12
W13
Reference Inlet Blocked Reference inlet blocked Check that nothing is blocking the inlet.
Reference inlet filter clogged
Contact INFICON service.
Sample Inlet Blocked Sample inlet blocked
If problems remains: Contact INFICON service.
Check that nothing is blocking the inlet
Sample inlet filter blocked Replace cigarette filter in sample inlet.
Blockage in accessory
Exhaust Blocked
O
2
sensor error
Internal filter blocked
Exhaust blocked
System error
Disconnect accessory.
If alarm disappears: Check accessory.
Disconnect accessories.
If alarm remains: Contact INFICON service.
Check that nothing is blocking the exhaust.
If alarm remains: Contact INFICON service.
Restart detector.
If problems remains: Contact INFICON service.
O
2
sensor error System error
H
H
2
2
S sensor error
S sensor error
CO sensor error
CO sensor error
CH
CH
CO
4
4
2
sensor error
sensor error
sensor error
System error
System error
System error
System error
System error
System error
System error
Restart detector.
If problems remains: Contact INFICON service.
Restart detector.
If problems remains: Contact INFICON service.
Restart detector.
If problems remains: Contact INFICON service.
Restart detector.
If problems remains: Contact INFICON service.
Restart detector.
If problems remains: Contact INFICON service.
Restart detector.
If problems remains: Contact INFICON service.
Restart detector.
If problems remains: Contact INFICON service.
Restart detector.
If problems remains: Contact INFICON service.
IRwin-Operating-instructions-mina66en1-15-(1904) 85 / 128
6 | Operation INFICON
No.
Message
W14 CO
2
W17 H
2
S STEL level reached, (STEL and limit shown,Confirm)
W18 H
2
W19
W20
W21
W22
sensor error
S LTEL level reached, (LTEL and limit shown,Confirm)
CO STEL level reached, (STEL and limit shown,Confirm)
CO LTEL level reached, (LTEL and limit shown,Confirm)
CO2 STEL level reached, (STEL and limit shown,Confirm)
CO2 LTEL level reached, (LTEL and limit shown,Confirm)
Possible error sources
System error
Short Term Exposure alarm for H alarm for H
2
S.
Long Term Exposure
2
S.
Short Term Exposure alarm for CO
Long Term Exposure alarm for CO
Short Term Exposure alarm for CO
2
Long Term Exposure alarm for CO
2
.
.
Remedy
Restart detector.
If problems remains: Contact INFICON service.
Normal function.
If warning reapperas without any gas present:
Recalibrate TOX mode.
If still reappearing: Contact INFICON service
Normal function.
If warning reapperas without any gas present:
Recalibrate TOX mode.
If still reappearing: Contact INFICON service
Normal function.
If warning reapperas without any gas present:
Recalibrate TOX mode.
If still reappearing: Contact INFICON service
Normal function.
If warning reapperas without any gas present:
Recalibrate TOX mode.
If still reappearing: Contact INFICON service
Normal function.
If warning reapperas without any gas present:
Recalibrate TOX mode.
If still reappearing: Contact INFICON service
Normal function.
If warning reapperas without any gas present:
Recalibrate TOX mode.
If still reappearing: Contact INFICON service
Error messages
E1 Main Pump error
E2
E3
E4
Sample Pump error
IR Lamp error
IR Lamp error
Exhaust blocked Check that nothing is blocking the exhaust.
If alarm remains: Contact INFICON service.
Reference inlet blocked Check that nothing is blocking the reference inlet.
Pump dirty or broken
If alarm remains: Contact INFICON service.
Contact INFICON service.
Sample pump not running Restart detector.
IR Lamp error
If problems remains: Contact INFICON service.
Restart detector.
IR lamp supply error
If problems remains: Contact INFICON service.
Restart detector.
If problems remains: Contact INFICON service.
86 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Operation | 6
No.
Message
E5 Battery level too low
(count down to power off)
E7 SnO
2
voltage error
Possible error sources
Battery low
Remedy
Switch off detector and charge battery
E8
E9
E10
Detector shuts down
Red LED and constant beep
O
2
Sensor bias lost
System error
System error
HW watchdog trigged
O2 sensor bias lost due to uncontrolled shut down.
Restart detector.
If problems remains: Contact INFICON service.
Restart detector.
If problems remains: Contact INFICON service.
Perform hardware reset.
Restart detector.
If problems remains: Contact INFICON service.
Check in Ex Tox mode that O
2
value is OK (close to 20.9%). If not, leave the detector on until O2 sensor shows 20 - 22% in air. This can take up to
12 hours.
Restart detector.
If problems remains: Contact INFICON service.
E11 “IR Lamp" has red cross in function test result screen.
E12 “Pump & Valve" test has red cross in function test result screen.
E13
E14
“Pump & Valve" test has red cross in function test result screen.
“Pump & Valve" test has red cross in function test result screen.
E15 "Gas Sensors" has red cross in function test result screen.
E16 "Gas Sensors" has red cross in function test result screen.
IR system error
Exhaust blocked
Pump dirty or broken
Restart detector.
If problems remains: Contact INFICON service.
Contact INFICON service.
Valve clogged Restart detector.
If problems remains: Contact INFICON service.
Restart detector.
System error
Exhaust blocked
If problems remains: Contact INFICON service.
Check that nothing is blocking the exhaust.
If alarm remains: Contact INFICON service.
Reference inlet blocked Check that nothing is blocking the reference inlet.
Pump dirty or broken
Sensor error or warm-up drift
Sensor error or warm-up drift
If alarm remains: Contact INFICON service.
Contact INFICON service.
Restart detector.
If problems remains: Contact INFICON service.
Restart detector.
If problems remains: Contact INFICON service.
IRwin-Operating-instructions-mina66en1-15-(1904) 87 / 128
6 | Operation INFICON
No.
Message
E17 "Gas Sensors" has red cross in function test result screen.
Possible error sources
Sensor error or warm-up drift
Remedy
Restart detector.
If problems remains: Contact INFICON service.
E18 "Buzzer" has red cross in function test result screen.
E23 "Gas Sensors" has red cross in function test result screen.
Buzzer "hole" blocked or filled with water
E19 "LED" has red cross in function test result screen.
E20
E21
E22
RAM test error (Power off)
ROM test error (Power off)
CRC error (power off)
LED broken
If problems remains: Contact INFICON service.
Contact INFICON service.
Memory damaged or test error
Restart detector.
Memory damaged or test error
If problems remains: Contact INFICON service.
Restart detector.
Memory damaged or test error
If problems remains: Contact INFICON service.
Restart detector.
Sensor error or warm-up drift
Dry and clean buzzer hole.
Restart detector.
If problems remains: Contact INFICON service.
Restart detector.
If problems remains: Contact INFICON service.
E24 No message given
E25 Surrounding temperature is outside of rated temperature.
Function might be affected
Service info only No action needed.
Instrument temperature is too high or to low
Explosion protection cannot be guaranteed.
Immediately move detector to cooler place or switch OFF if flammable gases could be present.
88 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Maintenance | 7
7 Maintenance
The gas inlet of the instrument, as well as the probe system include various filters.
These must be replaced at intervals specified in the maintenance plan. If the leak detector is used in dirty environments, more frequent replacements may be necessary.
NOTICE
The internal sample inlet filter can become clogged.
Running the leak detector with no probe connected can make the internal filters clogged. This can cause need for service.
7.1 Maintenance plan
Maintenance cycle daily weekly monthly
Additional information
Change the cigarette filter in the sample inlet
Maintenance of the Carpet Probe
Maintenance of the Carpet Probe
Maintenance of the Bar hole probe
7.2 Spare parts list
Spare part
Filter cartridge 0.45 um
Filter disk OD=8 mm
Dust filter, 8612
Order number
581-704
591-367
591-095
Cigarette Filter 120-pack 581-700
Bar Hole Rubber
Purpose
Carpet
Bell
Bell, Swan Neck and Flexi Bell
Hand Probe and Sample inlet
Seal for Bar Hole Probe
IRwin-Operating-instructions-mina66en1-15-(1904) 89 / 128
7 | Maintenance INFICON
7.3 General check of the probe system
The probe system used must be checked regularly for external damage and leaks.
1 Examine all parts of the probe system for external damages.
2 Check that the rubber grip of the Hand Probe is not broken or missing.
The rubber is certified anti-static and must be replaced by original INFICON spare if damaged or missing.
3 Connect the probe system to the instrument.
4 Block the inlet of the probe system.
ð If the error message "Blocked flow alarm" appears, the probe system is sufficiently leak-proof. When the leak detector does not show an error there is a leak in the system.
90 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Maintenance | 7
7.4 Maintenance of the Leak Detector
7.4.1 Seasonal shut-down
IRwin SXT is using a small amount of power when shut off. This allows the clock to continue running and the function of oxygen sensor to be maintained for faster warmup.
When you want to store the leak detector up to 3 months charge it fully.
When you want to store the leak detector more than 3 months select one of the following possibilities.
Connect the charging equipment and leave the charger on.
OR:
Charge the battery every 3 months as long as you store the leak detector.
7.4.2 Change the cigarette filter in the sample inlet
► Change the filter according to the maintenance plan, see Maintenance plan [ } 89].
Be careful when checking the cigarette filter in the inlet, make sure that it is not deformed, otherwise it will loose part of its function. Replace the filter if in doubt.
7.5 Maintenance of the Hand Probe
Replace the filter in the tip of the probe according to the maintenance plan, see
Maintenance plan.
Be careful when checking the cigarette filter in the inlet, make sure that it is not deformed, otherwise it will loose part of its function. Replace the filter if in doubt.
IRwin-Operating-instructions-mina66en1-15-(1904) 91 / 128
7 | Maintenance INFICON
7.6 Maintenance of the Carpet Probe
► Change the inline filter according to the maintenance plan, see Maintenance plan
92 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON
7.7 Maintenance of the Bell Probe
Maintenance | 7
Fig. 21: Flexi Bell
► Change the filter according to the maintenance plan, see Maintenance plan.
1 2 3 4
Fig. 22: Bell Probe
1
2
Bell
Paper filter
3
4
O-Ring
Filter housing
1 Screw off the top of the filter housing.
2 Remove the old paper filter and o-ring.
3 Place the o-ring around end cap on top of the new paper filter.
4 Place the filter and o-ring in the compartment of the filter housing.
5 Screw on the top tightly by hand.
6 Replace the bell probe.
IRwin-Operating-instructions-mina66en1-15-(1904) 93 / 128
7 | Maintenance
7.8 Maintenance of the Swan neck
INFICON
Fig. 23: Swan Neck
Change the filter according to the maintenance plan, see Maintenance plan.
See also
2
Maintenance of the Bell Probe [ } 93]
7.9 Maintenance of the Bar hole probe
Rubber needs to be replaced when worn. Contact INFICON.
Fig. 24: Bar Hole Probe
94 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Decommissioning the leak detector | 8
8 Decommissioning the leak detector
8.1 Sending in the leak detector
WARNING
Danger due to harmful substances
Contaminated equipment can endanger the health. The contamination declaration serves to protect all persons who come into contact with the equipment.
► Fill in the declaration of contamination completely.
1 Please do not hesitate to contact us and send a completed declaration of contamination before sending anything to us.
ð You will then receive a return number from us.
2 Use the original packaging when returning.
3 Before sending the instrument attach a copy of the completed contamination declaration. See below or on the homepage of INFICON.
IRwin-Operating-instructions-mina66en1-15-(1904) 95 / 128
8 | Decommissioning the leak detector INFICON
96 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Decommissioning the leak detector | 8
8.2 Disposing of the equipment
The leak detector can either be disposed of by the owner or be sent to the manufacturer. The leak detector contains materials that can be recycled. This option should be exercised to prevent waste and also to protect the environment.
During disposal, observe the environmental and safety regulations of your country.
Information for recycling.
The leak detetctor cannot be scrapped with the normal domestic waste.
IRwin-Operating-instructions-mina66en1-15-(1904) 97 / 128
9 | Certificates
9 Certificates
INFICON
Fig. 25: Certificates and permissions
98 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Certificates | 9
IRwin-Operating-instructions-mina66en1-15-(1904) 99 / 128
9 | Certificates INFICON
9.1 TÜV, Certificate for lower explosive limit (LEL)
100 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Certificates | 9
IRwin-Operating-instructions-mina66en1-15-(1904) 101 / 128
9 | Certificates INFICON
102 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Certificates | 9
IRwin-Operating-instructions-mina66en1-15-(1904) 103 / 128
9 | Certificates
9.2 Certificate for oxygen and toxic gases
INFICON
104 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Certificates | 9
IRwin-Operating-instructions-mina66en1-15-(1904) 105 / 128
9 | Certificates
9.3 Certificate for software for Ex/ExTox
INFICON
106 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Certificates | 9
9.4 ATEX certificate for explosion protection (IRwin
SXnn)
IRwin-Operating-instructions-mina66en1-15-(1904) 107 / 128
9 | Certificates
9.5 CSA (US/Canada) certificate for explosion protection (IRwin SXnn)
INFICON
108 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Certificates | 9
9.6 IECEx certificate for explosion protection (IRwin
SXnn)
IRwin-Operating-instructions-mina66en1-15-(1904) 109 / 128
9 | Certificates INFICON
110 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Certificates | 9
IRwin-Operating-instructions-mina66en1-15-(1904) 111 / 128
9 | Certificates
9.7 NEPSI (China) certificate for explosion protection
INFICON
112 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Certificates | 9
IRwin-Operating-instructions-mina66en1-15-(1904) 113 / 128
9 | Certificates INFICON
114 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Certificates | 9
IRwin-Operating-instructions-mina66en1-15-(1904) 115 / 128
9 | Certificates INFICON
116 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Certificates | 9
IRwin-Operating-instructions-mina66en1-15-(1904) 117 / 128
9 | Certificates
9.8 CE Declaration of Conformity
INFICON
118 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Certificates | 9
IRwin-Operating-instructions-mina66en1-15-(1904) 119 / 128
9 | Certificates INFICON
120 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Certificates | 9
IRwin-Operating-instructions-mina66en1-15-(1904) 121 / 128
9 | Certificates INFICON
122 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON Certificates | 9
IRwin-Operating-instructions-mina66en1-15-(1904) 123 / 128
Index
Index
A
Adjustment Factor
Alarm
Latching Function
Level alive signal
Ambient conditions
Application mode
Above Ground
Bar Hole
Confined Space
Ex
Ex Tox
Gas Purity
GC Ethane Analysis
House
Universal
Applications
Assembling
Bell Probe
Carpet Probe
Hand Probe
ATEX application
ATEX rating
B
Battery
Charging
Charging Adapter
Fast Charge
Bluetooth
C
Calibration
AbG Setup
AbG Test
124 / 128
INFICON
D
Dead band
Declaration of Contamination
Delete file
Description
Bar Hole Probe
Bell Probe
Carpet Probe
Flexi Bell
Function
Gases
Hand Probe
Probes
Swan Neck
Display
Function buttons
Icons
Navigation button
Operation Mode
Permission levels
Status
Documentation of inspections
E
Electrical data
Ethane Analysis
Determine Automatically
Recall information
F
Factory settings
Filter
Cigarette Filter
Inline Filter
Function buttons
Function test
IRwin-Operating-instructions-mina66en1-15-(1904)
INFICON
G
Gas
Select Gas
I
Installation
Measuring system
Intended use
IR Ethane Analysis
L
Language
M
Maintenance
Bar Hole Probe
Bell probe
Hand probe
Swan Neck
Measurement Data Recording
N
Navigation buttons
O
Settings
P
Permission levels
Personnel qualifications
Physical data
Probes
R
Return
S
Scope of delivery
Screen timeout
Self-test
Settings
Measurement
Spare parts statistics
Status
Status display
Status indicators
Switch gas
Switch OFF
T
Technical Data
Ambient Conditions
Electrical Data
Mechanical Data
Physical Data
Time local time
Type certificate
U
Universal Mode
W
Warning and error messages
Warning Ex
Index
IRwin-Operating-instructions-mina66en1-15-(1904) 125 / 128
Index INFICON
126 / 128 IRwin-Operating-instructions-mina66en1-15-(1904)
advertisement
Related manuals
advertisement
Table of contents
- 7 1 About these instructions
- 7 1.1 Target groups
- 7 1.2 Warnings
- 8 2 Safety
- 8 2.1 Intended use
- 9 2.2 Special conditions for safe use
- 9 2.2.1 Intrinsic Safety (Ex protection)
- 10 2.2.2 Certified Gas Measurement
- 11 2.3 Owner/supervisor responsibilities
- 12 2.4 Duties of the operator
- 12 2.5 Safe operation
- 13 3 Scope of delivery
- 14 4 Description
- 14 4.1 Function
- 15 4.2 Design of the leak detector
- 16 4.3 Probes
- 19 4.4 Display
- 22 4.5 Technical data
- 24 4.5.1 EX certification (intrinsic safety)
- 24 4.5.2 Sensors
- 32 4.6 Factory settings
- 34 4.7 Concentration of calibration and test gases
- 35 5 Getting started
- 35 5.1 Charging the battery
- 38 5.2 Assembling the probe system
- 40 5.3 Transfer data to computer
- 40 5.3.1 Install software for data transfer
- 40 5.3.2 Pairing the leak detector with computer via Bluetooth interface
- 41 5.3.3 Transfer log files from leak detector to computer
- 41 5.3.4 Real-time streaming from leak detector to computer
- 43 6 Operation
- 43 6.1 Switch ON
- 44 6.2 Initial setup
- 44 6.2.1 Select available operating modes
- 44 6.2.2 Adjust the local time
- 44 6.2.3 Setting the language
- 45 6.2.4 Change permission level
- 45 6.2.5 PPM Adjustment Factor
- 45 6.2.6 Select Gas (Ex)
- 46 6.2.7 100% CH4 LEL conc
- 46 6.2.8 Universal Mode
- 46 6.2.9 AbG Cal Setup
- 46 6.2.10 AbG Fnc Test conc
- 47 6.2.11 Screensaver timeout
- 48 6.2.12 Configure data logging
- 48 6.2.13 Enable dead band suppression (IRwin SXnn only)
- 48 6.2.14 Setting the Function Test Interval
- 49 6.2.15 Calibration
- 51 6.2.15.1 Standard calibration procedures
- 52 6.2.15.2 Setting of calibration gas concentrations
- 53 6.2.15.3 Adjust calibration reminder
- 54 6.3 Operating modes
- 54 6.3.1 Select operating mode
- 54 6.3.2 Operating mode “Above Ground
- 56 6.3.3 Operating mode "Universal
- 58 6.3.4 Operating mode “Bar Hole
- 59 Automatic “Bar Hole” testing
- 60 Setting parameters for “Auto Bar Hole
- 61 6.3.5 Operating mode "GC Ethane Analysis
- 65 6.3.6 Operating mode “Confined Space
- 67 6.3.7 Operating mode "House
- 68 6.3.8 Operating mode "Gas Purity
- 70 6.3.9 Operating mode "Ex
- 73 6.3.10 Operating mode "Ex Tox
- 75 6.3.10.1 Change gas type for %LEL measurement
- 76 6.3.11 IR Ethane Analysis and compensation
- 76 6.3.11.1 Entering ethane concentration manually
- 77 6.3.11.2 Setting reminder for IR Ethane Analysis
- 77 6.3.11.3 Performing IR Ethane Analysis after automatic reminder
- 78 6.3.11.4 Starting IR Ethane Analysis manually
- 79 6.3.11.5 Viewing IR Ethane Analysis info
- 79 6.3.12 Set limits and alarm levels
- 80 6.3.13 Set the brightness of the display
- 80 6.3.14 Performing function tests
- 81 6.4 Measuring
- 82 6.5 Perform self test
- 82 6.6 Viewing Status menu
- 82 6.7 Logging data to file
- 83 6.8 Viewing logfile list
- 83 6.9 Information
- 83 6.9.1 Retrieve information and statistics
- 83 6.9.2 Viewing the Service Screen
- 83 6.10 Activating Auto-Rotate Screen
- 84 6.11 Switching OFF
- 84 6.12 Warnings and error messages
- 89 7 Maintenance
- 89 7.1 Maintenance plan
- 89 7.2 Spare parts list
- 90 7.3 General check of the probe system
- 91 7.4 Maintenance of the Leak Detector
- 91 7.4.1 Seasonal shut-down
- 91 7.4.2 Change the cigarette filter in the sample inlet
- 91 7.5 Maintenance of the Hand Probe
- 92 7.6 Maintenance of the Carpet Probe
- 93 7.7 Maintenance of the Bell Probe
- 94 7.8 Maintenance of the Swan neck
- 94 7.9 Maintenance of the Bar hole probe
- 95 8 Decommissioning the leak detector
- 95 8.1 Sending in the leak detector
- 97 8.2 Disposing of the equipment
- 98 9 Certificates
- 100 9.1 TÜV, Certificate for lower explosive limit (LEL)
- 104 9.2 Certificate for oxygen and toxic gases
- 106 9.3 Certificate for software for Ex/ExTox
- 107 9.4 ATEX certificate for explosion protection (IRwin SXnn)
- 108 9.5 CSA (US/Canada) certificate for explosion protection (IRwin SXnn)
- 109 9.6 IECEx certificate for explosion protection (IRwin SXnn)
- 112 9.7 NEPSI (China) certificate for explosion protection
- 118 9.8 CE Declaration of Conformity
- 124 Index