Below you will find brief information for blood pressure monitor HEM-7211. The OMRON HEM-7211 is a compact, fully automatic blood pressure monitor, operating on the oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. Equipped with Dual Check System and Cuff Wrapping Guide, HEM-7211 provides accurate and anxiety-free blood pressure measurement everyday.
Advertisement
Advertisement
Automatic Blood Pressure Monitor
Model HEM-7211
Instruction Manual
All for Healthcare
HEM-7211-C1_A_M.book Page 2 Tuesday, February 23, 2010 1:15 PM
Contents
Before using the unit
Introduction ................................................................................ 3
Important Safety Information .................................................... 4
1. Overview ............................................................................... 7
2. Preparation ......................................................................... 10
2.1 Installing/Replacing the Batteries ..............................................10
2.2 Setting the Date and Time .........................................................11
Operating instructions
3. Using the Unit ..................................................................... 12
3.1 Applying the Arm Cuff ................................................................12
3.2 How to Sit Correctly ...................................................................14
3.3 Taking a Reading .......................................................................15
3.4 Using the Memory Function .......................................................18
4. Troubleshooting and Maintenance .................................. 20
4.1 The Icons and Error Messages ..................................................20
4.2
Troubleshooting .........................................................................22
4.3
Maintenance ..............................................................................24
4.4
Storage ......................................................................................25
Care and maintenance
5. Optional Parts .................................................................... 26
6. Technical Data .................................................................... 27
2
HEM-7211-C1_A_M.book Page 3 Tuesday, February 23, 2010 1:15 PM
Before using the unit
Introduction
Introduction
Thank you for purchasing the OMRON HEM-7211 Automatic Blood
Pressure Monitor.
The OMRON HEM-7211 is a compact, fully automatic blood pressure monitor, operating on the oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. Equipped with Dual
Check System and Cuff Wrapping Guide, HEM-7211 provides accurate and anxiety-free blood pressure measurement everyday. For comfortable controlled inflation without the need of pressure presetting or re-inflation the devices uses its advanced “IntelliSense” technology.
Intended Use
This product is designed to measure the blood pressure and pulse rate of people within the range of the designated arm cuff, following the instructions in this instruction manual.
It is mainly designed for general household use. Please read the
Important Safety Information in this instruction manual before using the unit.
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure,
CONSULT YOUR DOCTOR.
3
HEM-7211-C1_A_M.book Page 4 Tuesday, February 23, 2010 1:15 PM
Important Safety Information
Important Safety Information
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or arteriosclerosis.
Please read this section carefully before using the unit.
Warning:
• Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and selftreatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a doctor before using the unit, as cuff inflation can cause internal bleeding.
(Optional AC Adapter Usage)
• Never plug in or unplug the power cord from the electric outlet with wet hands.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.
4
HEM-7211-C1_A_M.book Page 5 Tuesday, February 23, 2010 1:15 PM
Important Safety Information
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result in incorrect measurement results.
• Make sure that the air tube is not wrapped around other parts of your body when taking measurements. This could result in injury when the air pressure in the air tube is increased.
• Do not leave the cuff wrapped on the arm if taking measurements during the night. This could result in injury.
• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
• Do not operate the unit in a moving vehicle (car, airplane).
• To inflate the cuff manually, refer to Section 3.3. If the cuff is over inflated, it can cause internal bleeding.
(Optional AC Adapter Usage)
• Use only the original AC adapter designed for this unit. Use of unsupported adapters may damage and/or may be hazardous to the unit.
• Plug the AC adapter into the appropriate voltage outlet. Do not use a multipletap.
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off the power and unplug the power cord immediately.
5
HEM-7211-C1_A_M.book Page 6 Tuesday, February 23, 2010 1:15 PM
Important Safety Information
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with plenty of clean water.
• Use only four “AA” alkaline or manganese batteries with this unit. Do not use other types of batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace all four batteries at the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol or caffeine, smoking, exercising or eating.
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull on the air plug at the connection with the main unit not the tube itself.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Read and follow the "Important information regarding Electro Magnetic
Compatibility (EMC)" in the Technical Data Section.
• Read and follow the "Correct Disposal of This Product" in the Technical
Data Section when disposing of the device and any used accessories or optional parts.
6
HEM-7211-C1_A_M.book Page 7 Tuesday, February 23, 2010 1:15 PM
1. Overview
Main Unit:
A
G
1. Overview
F
E
B
C
D
A. Display
B. Memory button
C. START/STOP button
D. Up/Down ( / ) buttons
E. Date/Time setting ( ) button
F. Air Jack
G. Battery compartment
H. AC adapter jack (for optional
AC adapter)
7
HEM-7211-C1_A_M.book Page 8 Tuesday, February 23, 2010 1:15 PM
1. Overview
Arm Cuff:
I J K
L
I. Arm cuff
(Medium cuff: arm circumference 22-32 cm)
J. Marker
K. Air tube
L. Air plug
Storage case:
8
HEM-7211-C1_A_M.book Page 9 Tuesday, February 23, 2010 1:15 PM
Display:
P
Q
R
M
N
O
S
T
U
V
W
X
Y
Z
1. Overview
M. Date/Time display
N. Dual Check System
O. Systolic blood pressure
P. Diastolic blood pressure
Q. Heartbeat symbol
(Flashes during measurement)
R. Battery low symbol
S. Average value symbol
T. Memory symbol
U. Irregular heartbeat symbol
V. Movement error symbol
W. Blood pressure level indicator
X. Deflation symbol
Y. Cuff wrapping guide
Z. Pulse display and Memory number
9
HEM-7211-C1_A_M.book Page 10 Tuesday, February 23, 2010 1:15 PM
2. Preparation
2. Preparation
2.1
Installing/Replacing the Batteries
1.
Remove the battery cover.
2.
Insert four “AA” batteries as indicated in the battery compartment and then replace the battery cover.
Notes:
• If the battery low symbol
( ) appears on the display, turn the unit off then replace all batteries at the same time.
• The measurement values continue to be stored in memory even after the batteries are replaced.
Disposal of used batteries should be carried out in accordance with the national regulations for the disposal of batteries.
10
HEM-7211-C1_A_M.book Page 11 Tuesday, February 23, 2010 1:15 PM
Operating instructions
2.2
Setting the Date and Time
2. Preparation
1.
Press the button.
2.
Set the monitor to the correct date and time before taking a measurement for the first time.
1)
Push or buttons to change.
2)
Push the button to confirm.
Next setting appears.
Year Month Day Hour Minute
3.
Press the START/STOP button to store the setting.
Notes:
• If the batteries have been removed for 30 seconds or more, the date and time setting will need to be reset.
• If the date and time are not set, “-:--” appears during or after measurement.
11
HEM-7211-C1_A_M.book Page 12 Tuesday, February 23, 2010 1:15 PM
3. Using the Unit
3. Using the Unit
3.1
Applying the Arm Cuff
Remove tight-fitting clothing or tight rolled up sleeve from your upper arm.
Do not place the cuff over thick clothes.
1.
Insert the air plug into the air jack.
2.
Put your arm through the cuff loop.
12
3.
Position the arm correctly.
HEM-7211-C1_A_M.book Page 13 Tuesday, February 23, 2010 1:15 PM
3. Using the Unit
The bottom edge of the cuff should be 1 to 2 cm above the elbow.
Marker (arrow under tube) is centred on the middle of your inner arm.
Close the fabric fastener FIRMLY.
1 to 2 cm
Notes:
• When you take a measurement on the right arm, air tube will be at the side of your elbow. Be careful not to rest your arm on the air tube.
• The blood pressure can differ between the right arm and the left arm, and therefore also the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between the two arms differ substantially, please check with your doctor which arm to use for your measurement.
13
HEM-7211-C1_A_M.book Page 14 Tuesday, February 23, 2010 1:15 PM
3. Using the Unit
3.2
How to Sit Correctly
To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. No bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement.
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight.
• The cuff should be at the same level as your heart.
14
HEM-7211-C1_A_M.book Page 15 Tuesday, February 23, 2010 1:15 PM
3. Using the Unit
3.3
Taking a Reading
Notes:
• To cancel a measurement, press the START/STOP button to turn the unit off and to release the air in the arm cuff.
• Remain still while taking a measurement.
1.
Press the START/STOP button.
The cuff will start to inflate automatically.
START INFLATE DEFLATE END COMPLETE
Dual
Check
System
Cuff Wrapping Guide
If your systolic pressure is more than
220 mmHg
After the cuff starts to inflate, press and hold the
START/STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
Notes:
• The monitor will not inflate above 299 mmHg.
• Do not apply more pressure than necessary.
Dual Check System
Press the START/STOP button, then the Dual Check System lamp lights and the unit starts detecting a device error. If the unit is accurate and functioning correctly, the Dual Check
System lamp remains lit during the measurement. If an error is detected, the Dual Check System lamp flashes and “Er” will appear on the display.
15
HEM-7211-C1_A_M.book Page 16 Tuesday, February 23, 2010 1:15 PM
3. Using the Unit
Cuff Wrapping Guide
The Cuff Wrapping Guide is a unique feature that indicates if the cuff is not wrapped tightly enough around the arm. Even when the is displayed, a blood pressure reading will be taken.
Note:
This reading is NOT reliable due to the incorrect wrapping
of the cuff. Please wrap the cuff again, taking care to wrap
it correctly and take the measurement again.
When the is displayed, the cuff is correctly wrapped
tightly enough on the arm and the reading is accurate and
reliable.
2.
Remove the arm cuff.
3.
Press the START/STOP button to turn the monitor off.
The monitor automatically stores the measurement in its memory.
It will automatically turn off after 2 minutes.
Important:
• Recent research suggests that the following values can be used as a guide to high blood pressure for measurements taken at home.
Systolic Blood Pressure Above 135 mmHg
Diastolic Blood Pressure Above 85 mmHg
This criteria is for home blood pressure measurement.
16
HEM-7211-C1_A_M.book Page 17 Tuesday, February 23, 2010 1:15 PM
3. Using the Unit
• Your blood pressure monitor includes an irregular heartbeat feature. Irregular heartbeats can influence the results of the measurement. The irregular heartbeat algorithm automatically determines if the measurement is usable or needs to be repeated. If the measurement results are affected by irregular heartbeats but the result is valid, the result is shown together with the irregular heartbeat symbol ( ). If the irregular heartbeats cause the measurement to be invalid, no result is shown. If the irregular heartbeat symbol ( ) is shown after you have taken a measurement, repeat the measurement. If the irregular heartbeat symbol ( ) is shown frequently, please make your doctor aware of it.
• If you move during measurement, the movement error symbol ( ) will appear on the display.
Keep still and repeat the measurement.
Note:
Wait 2-3 minutes before taking another blood pressure measurement. Waiting between readings allows the arteries to return to the condition prior to taking the blood pressure measurement.
Warning:
Self-diagnosis of measured results and treatment are dangerous.
Please follow the instructions of your doctor.
17
HEM-7211-C1_A_M.book Page 18 Tuesday, February 23, 2010 1:15 PM
3. Using the Unit
3.4
Using the Memory Function
The monitor automatically stores the result up to 90 sets. It can also calculate an average reading based on the measurements from the last three readings taken within 10 minutes. If there are only two readings in memory for that period, the average will be based on two readings. If there is one reading in memory for that period, the average will be based on one reading.
Notes:
• If the memory is full, the monitor will delete the oldest readings.
• When viewing the reading taken without setting the date and time, “-:--” is displayed instead of the date and time.
To View the Average Value
1.
Press the Memory button.
Note:
If there are no measurements results stored in the memory, the screen to the right is displayed.
18
HEM-7211-C1_A_M.book Page 19 Tuesday, February 23, 2010 1:15 PM
3. Using the Unit
To View the Readings Stored in Memory
1.
Press the button, while the average value is displayed.
The Memory number appears for a second before the pulse rate is displayed. The newest set is numbered “1”.
Alternating date/ display
2.
Press the or button to view the readings stored in memory.
: To the older readings
: To the more recent readings
To delete all the values stored in memory
When the memory symbol ( ) appears, first press the
Memory button. Then while holding it down, press the
START/STOP button simultaneously for about 2-3 seconds.
Note:
You cannot partially delete the values stored in the memory.
19
HEM-7211-C1_A_M.book Page 20 Tuesday, February 23, 2010 1:15 PM
4. Troubleshooting and Maintenance
4. Troubleshooting and Maintenance
4.1
The Icons and Error Messages
Error Display Cause
Irregular heartbeats are detected.
Movement during measurement.
Remedy
Remove the arm cuff. Wait 2-3 minutes and then take another measurement.
Repeat the steps in section 3.3. If this error continues to appear, contact your doctor.
Carefully read and repeat the steps in section 3.3.
Apply the arm cuff correctly.
Refer to section 3.1.
Blink
The batteries are low.
Lit
The batteries are exhausted.
ones ahead of time.
Refer to section 2.1.
You should replace them with new ones at once.
Refer to section 2.1.
20
HEM-7211-C1_A_M.book Page 21 Tuesday, February 23, 2010 1:15 PM
Error Display
4. Troubleshooting and Maintenance
Cause
Air plug disconnected.
Arm cuff not applied correctly.
Air is leaking from the arm cuff.
Cuff over inflated.
The arm cuff was inflated above 299 mmHg when inflating the cuff manually.
Movement during measurement.
Clothing is interfering with the arm cuff.
Remedy
Insert the plug securely.
Refer to section 3.1.
Apply the arm cuff correctly.
Refer to section 3.1.
Replace the cuff with the new one.
Refer to Chapter 5.
Repeat measurement. Remain still and do not talk during measurement.
Refer to section 3.3.
If “E2” appears repeatedly, inflate the cuff manually until it is 30 to 40 mmHg above your previous measurement result.
Refer to section 3.3.
Do not inflate the cuff above 299 mmHg.
Refer to section 3.3.
Repeat measurement. Remain still and do not talk during measurement.
Refer to section 3.3.
Remove any clothing interfering with the arm cuff.
Refer to section 3.1.
Device error.
Contact your OMRON retail outlet or distributor.
Note:
The irregular heartbeat symbol ( ) may also be displayed with error messages.
21
HEM-7211-C1_A_M.book Page 22 Tuesday, February 23, 2010 1:15 PM
4. Troubleshooting and Maintenance
4.2
Troubleshooting soon.
Problem
The reading is extremely low (or high).
Arm cuff pressure does not rise.
Arm cuff deflates too
Cannot measure or readings are too low or too high.
Nothing happens when you press the buttons.
Cause
Arm cuff not applied correctly.
Movement or talking during measurement.
Clothing is interfering with the arm cuff.
The air tube is not securely connected into the main unit.
Air is leaking from the arm cuff.
The arm cuff is loose.
The arm cuff has not been inflated sufficiently.
The batteries are empty.
The batteries have been inserted incorrectly.
Remedy
Apply the arm cuff correctly.
Refer to section 3.1.
Remain still and do not talk during measurement.
Refer to section 3.3.
Remove any clothing interfering with the arm cuff.
Refer to section 3.1.
Make sure that the air tube is connected securely.
Refer to section 3.1.
Replace the arm cuff with a new one.
Refer to Chapter 5.
Apply the cuff correctly so that it is firmly wrapped around the arm.
Refer to section 3.1.
Inflate the cuff so that it is
30 to 40 mmHg above your previous measurement result.
Refer to section 3.3.
Replace the batteries with new ones.
Refer to section 2.1.
Insert the batteries with the correct (+/ -) polarity.
Refer to section 2.1.
22
HEM-7211-C1_A_M.book Page 23 Tuesday, February 23, 2010 1:15 PM
Problem
Other problems.
4. Troubleshooting and Maintenance
Cause Remedy
• Press the START/STOP button and repeat measurement.
• If the problem continues, try replacing the batteries with new ones.
If this still does not solve the problem, contact your local authorised OMRON distributor or dealer.
23
HEM-7211-C1_A_M.book Page 24 Tuesday, February 23, 2010 1:15 PM
4. Troubleshooting and Maintenance
4.3
Maintenance
To protect your unit from damage, please observe the following:
• Do not subject the main unit and the cuff to extreme temperatures, humidity, moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations (for example, dropping the unit on the floor).
• Do not use volatile liquids to clean the main unit.
• Do not wash the arm cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to clean the arm cuff.
• Do not carry out repairs of any kind yourself. If a defect occurs, consult your local authorised OMRON distributor or dealer.
24
• The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and soap to clean the arm cuff.
HEM-7211-C1_A_M.book Page 25 Tuesday, February 23, 2010 1:15 PM
4. Troubleshooting and Maintenance
Calibration and Service
• The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service life.
• It is generally recommended to have the unit inspected every two years to ensure correct functioning and accuracy. Please consult your local authorised OMRON distributor or dealer.
4.4
Storage
Keep the unit in its storage case when not in use.
1.
Unplug the air plug from the air jack.
2.
Gently fold the air tube into the arm cuff.
Note:
Do not bend the air tube excessively.
3.
Place the arm cuff and main unit in the storage case.
Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will be at an angle.
Care and maintenance
25
HEM-7211-C1_A_M.book Page 26 Tuesday, February 23, 2010 1:15 PM
5. Optional Parts
5. Optional Parts
Medium Arm Cuff
Arm circumference
22 - 32 cm
Large Arm Cuff
Arm circumference
32 - 42 cm
Small Arm Cuff
Arm circumference
17 - 22 cm
AC Adapter
S
Using the Optional AC Adapter
1.
Insert the AC adapter plug into the AC adapter jack on the rear side of the main unit.
2.
Plug the AC adapter into an electrical outlet.
To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug from the main unit.
26
9515336-9
HEM-7211-C1_A_M_1.fm Page 27 Wednesday, June 2, 2010 10:01 AM
6. Technical Data
6. Technical Data
Product Description
Model
Display
Measurement Method
Measurement Range
Accuracy
Inflation
Deflation
Memory
Rating
Power Source
Battery Life
Automatic Blood Pressure Monitor
HEM-7211 (HEM-7211-C1(VG))
LCD Digital Display
Oscillometric method
Pressure: 0 to 299 mmHg
Pulse: 40 to 180/min.
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Fuzzy-logic controlled by electric pump
Automatic pressure release valve
90 Measurements with date and time
DC6V 4W
(optional), INPUT AC100-240V 50/60Hz
S 9515336-9
0.12A)
Capacity of new alkaline batteries is approx. 1500 measurements
Applied Part
= Type B
Protection Against Electric
Shock
Internally powered ME equipment (When using only the batteries)
= Class II ME equipment (Optional AC adapter)
Operating temperature/
Humidity
Storage temperature/ Humidity/
Air pressure
Console Weight
Cuff Weight
Outer Dimensions
Cuff Dimensions
Cuff Material
Package Content
+10°C to +40°C / Maximum: 30 to 85% RH
-20°C to +60°C / Maximum: 10 to 95% RH / 700-1060 hPa
Approximately 380g without batteries
Approximately 120g
Approximately 123 (w) mm × 85 (h) mm × 158 (l) mm
Approximately 146 mm × 446 mm (Medium cuff: arm circumference 22 to 32 cm)
Nylon and polyester
Main unit, cuff, instruction manual, storage case, battery set
Note:
Subject to technical modification without prior notice.
27
HEM-7211-C1_A_M.book Page 28 Tuesday, February 23, 2010 1:15 PM
6. Technical Data
• This device fulfils the provisions of EC directive 93/42/EEC (Medical
Device Directive).
• This blood pressure monitor is designed according to the European
Standard EN1060, Noninvasive sphygmomanometers Part 1:
General Requirements and Part 3: Supplementary requirements for electromechanical blood pressure measuring systems.
• This OMRON product is produced under the strict quality system of
OMRON HEALTHCARE Co. Ltd., Japan. The Core component for
OMRON blood pressure monitors, which is the Pressure Sensor, is produced in Japan.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices.
Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the EN60601-1-2:2007 standard has been implemented. This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate strong electrical or electromagnetic fields, near the medical device. This may result in incorrect operation of the unit and create a potentially unsafe situation. Recommendation is to keep a minimum distance of 7 m.
Verify correct operation of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at OMRON HEALTHCARE
EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it should not be disposed of, with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this product from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can return this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial wastes for disposal.
28
HEM- 7211- C1_A_M03_100113. pdf
HEM-7211-C1_A_M_CS.fm 1
ページ 2010年2月19日 金曜日 午後8時3分
All for Healthcare
自动血压计
型号
HEM-7211
使用说明书
HEM-7211-C1_A_M_CS.fm 2 ページ 2010年2月19日 金曜日 午後8時3分
目录:
使用血压计前
简介 .................................................... 3
重要安全注意事项 ........................................ 4
1. 概要 ................................................ 7
2. 准备 ............................................... 10
2.1 安装 / 更换电池 ......................................... 10
2.2 设定日期和时间 ......................................... 11
操作说明
3. 使用血压计 ......................................... 12
3.1 缠绕臂带 ............................................... 12
3.2 如何保持坐姿正确 ....................................... 14
3.3 测量 ................................................... 15
3.4 使用记忆读取功能 ....................................... 18
4. 疑难排解和维修 ..................................... 20
4.1 标志与错误指示符 ....................................... 20
4.2 疑难排解 ............................................... 22
4.3 维修 ................................................... 24
4.4 储存 ................................................... 25
保养和维修
5. 选件 ............................................... 26
6. 技术资料 ........................................... 27
2
HEM-7211-C1_A_M_CS.fm 3 ページ 2010年2月19日 金曜日 午後8時3分
使用血压计前
简介
感谢您购买
OMRON HEM-7211 自动血压计。
简介
OMRON HEM-7211 是一款全自动血压计,根据示波法原理操作。本
产品可方便快捷地测量血压与脉搏速率。
HEM-7211 配备有双检测系
统 (
Dual Check System) 和臂带缠绕指示 (Cuff Wrapping Guide),可
让您每天都能测得准确血压,绝对放心。本设备利用先进的
IntelliSense 高智能 技术,既可舒适控制充气,又可免于预设压力或重
新充气。
使用本机前请先仔细阅读本说明书。
请妥善保管本说明书以备将来参考。
有关用户自身血压的具体详情,请咨询医生。
3
HEM-7211-C1_A_M_CS.fm 4 ページ 2010年2月26日 金曜日 午後4時30分
重要安全注意事项
重要安全注意事项
孕妇或经诊断为心律不齐或动脉硬化的患者,请在使用本产品前咨询医
生。
使用本产品之前请先细读本节内容。
警告:
• 表示潜在的危险情形,如未避免此类情形,可能导致死亡或严重受伤。
( 一般用途 )
• 始终咨询医生。依测量值自我诊断及自我治疗非常危险。
• 由于臂带充气后会导致内出血,因此血液流动存在严重问题或血液异常的患者
在使用本机前应咨询医生。
( 使用交流适配器 选件 )
• 请勿用湿手从插座插拔电源线。
( 电池使用 )
• 若电池溶液进入眼睛,请立刻用大量清水冲洗,并马上就诊。
注意:
• 表示潜在的危险情形,如未避免此类情形,可能导致用户或患者轻微或中等程
度的身体损伤,或损坏设备或其它财产。
4
HEM-7211-C1_A_M_CS.fm 5 ページ 2010年2月19日 金曜日 午後8時3分
重要安全注意事项
( 一般用途 )
• 本机不能用于婴儿或无法正确表达自己意愿的人士。
• 请勿用于测量血压以外的任何目的。
• 请仅使用本装置认可的臂带。使用其它臂带可能导致测量结果不准确。
• 请确保气管没有缠绕在您身体的其他部位。如若气管内气压升高,可能会导致
受伤。
• 夜晚测量时,请勿让臂带一直缠绕在手臂上。这可能导致受伤。
• 请勿对臂带充气超过 299 mmHg。
• 请勿在血压计附近使用手机或其他发射电磁场的装置。这可能导致血压计操作
出错。
• 请勿在行动的交通工具 (汽车、飞机)上使用血压计。
• 要对臂带手动充气,请参阅第 3.3 节。如果臂带充气过度,可导致内出血。
( 使用交流适配器 选件 )
• 请仅使用专为本装置所设计的原装交流适配器 。使用不受支持的适配器可能
对本装置造成损坏和 / 或危险。
• 将交流适配器插入适宜的电压插座。请勿使用多分接头插座。
• 如果本装置或电源线损坏,请勿使用交流适配器。立即关闭电源并拔下电源
线。
5
HEM-7211-C1_A_M_CS.fm 6 ページ 2010年2月19日 金曜日 午後8時3分
重要安全注意事项
( 电池使用 )
• 若电池溶液滴在皮肤或衣服上,请立刻用大量清水冲洗。
• 本机只使用四节 “AA”碱性电池或锰电池。切勿使用其它类型的电池。
• 切勿将电池电极装反。
• 及时用新的电池更换旧电池。同时将四个新电池全部装入。
• 如果三个月或更长时间内不使用血压计,请取出电池。
• 新旧电池不要混合使用。
一般注意事项
• 请勿使血压计受到猛烈震动或晃动;不要摔扔血压计及臂带。
• 请勿在沐浴、饮用含酒精或咖啡因饮品、吸烟、运动或进食后测量。
• 请勿用力弯曲臂带或过度拉伸气管。
• 卸除气管时,拔下与主机相连的气塞而不是气管本身。
• 臂带尚未缠绕于手臂上时,请勿给臂带充气。
6
HEM-7211-C1_A_M_CS.fm 7 ページ 2010年2月19日 金曜日 午後8時3分
1. 概要
主机:
A
G
1. 概要
F
E
B
C
D
A. 显示屏
B. “记忆”按钮
C. START/STOP “加压 / 停止”
按钮
D. 上 / 下 ( / ) 按钮
E. 日期 / 时间设定 ( ) 按钮
F. 充气孔
G. 电池仓
H. 交流适配器插孔
(用于交流适配器选件)
7
HEM-7211-C1_A_M_CS.fm 8 ページ 2010年2月19日 金曜日 午後8時3分
1. 概要
臂带:
I J K
储存盒:
I. 臂带
(中型臂带:手臂周长
22-32 厘米)
J. 标记
K. 气管
L. 气塞
L
8
HEM-7211-C1_A_M_CS.fm 9 ページ 2010年2月19日 金曜日 午後8時3分
显示屏:
P
Q
R
M
N
O
U
V
W
S
T
X
Y
Z
1. 概要
M. 日期 / 时间显示
N. 双检测系统 (Dual Check
System)
O. 收缩压
P. 舒张压
Q. 心律符号
(测量时闪烁)
R. 电池电量不足符号
S. 平均数值符号
T. 记忆符号
U. 非正规心律符号
V. 移动错误符号
W. 血压水平指示器
X. 排气符号
Y. 臂带缠绕指示 (Cuff Wrapping
Guide)
Z. 脉搏显示与记忆数位
9
HEM-7211-C1_A_M_CS.fm 10 ページ 2010年2月19日 金曜日 午後8時3分
2. 准备
2. 准备
2.1
安装 / 更换电池
1. 取下电池盖。
2. 按电池仓上标示的方向放入
四节 “
AA”电池,然后装回
电池盖。
注意:
•
如果显示屏出现电池电量不
足符号 ( ),请关闭本装
置并同时更换所有电池。
• 更换电池后,测量值仍会储存在记忆中。
旧电池的弃置应遵循国内对废旧电池的弃置规定。
10
HEM-7211-C1_A_M_CS.fm 11 ページ 2010年3月1日 月曜日 午後6時47分
操作说明
2.2
设定日期和时间
1. 按 按钮。
2. 第一次测量前,请将血压计设定到正确的日期及时间。
1)
按 或 按钮更改设定值。
2)
按 按钮确认。出现下一个设定。
2. 准备
年 月 日 时 分
3. 按 START/STOP “加压/停止”按钮储存设定。
注意:
•
•
如果取出电池时间超过
30 秒,则需重新设定日期和时间。
如果未设定日期和时间, “
-:--”
将在测量中和测量后显示在
显示屏上。
11
HEM-7211-C1_A_M_CS.fm 12 ページ 2010年3月1日 月曜日 午後6時47分
3. 使用血压计
3. 使用血压计
3.1
缠绕臂带
将紧身衣物从手臂脱下,或卷起衣袖。
请勿穿着厚衣服进行测量。
1. 将气塞插入充气孔中。
2. 将手臂穿过臂带回环。
3. 正确摆放手臂。
12
HEM-7211-C1_A_M_CS.fm 13 ページ 2010年2月19日 金曜日 午後8時3分
臂带下边缘应高出肘部
1 - 2 厘米。
标记 (气管下的箭头)应处于内臂的中央位置。
紧紧地合上黏扣带。
3. 使用血压计
1 至 2 厘米
注意:
•
用右臂进行测量时,气管应位于肘
部一侧。小心不要将手臂压在气管
上。
•
右臂和左臂上的血压可能各不相
同,因此测得的血压值可能也有所
差异。
OMRON 建议在测量时始终测量同一手臂。如果两臂
的测量值存在明显的差异,请咨询医生应使用哪个手臂进行测
量。
13
HEM-7211-C1_A_M_CS.fm 14 ページ 2010年2月19日 金曜日 午後8時3分
3. 使用血压计
3.2
如何保持坐姿正确
测量时,您应在舒适的室温环境下放松、端
坐。测量前
30 分钟内,不要沐浴、饮用含酒
精或咖啡因饮品、吸烟、锻炼或进食。
•
•
坐在椅子上,双脚平放于地面。
正襟危坐,背部笔直。
• 保持臂带与心脏同一水平。
14
HEM-7211-C1_A_M_CS.fm 15 ページ 2010年2月19日 金曜日 午後8時3分
3. 使用血压计
3.3
测量
注意:
•
要取消测量,按
START/STOP “加压/停止”按钮关闭装
置,释放掉臂带中的空气。
•
测量时保持静坐。
1. 按 START/STOP “加压/停止”按钮。
臂带开始自动充气。
加压 充气 排气 结束 完成
双检测
系统
臂带缠绕指示
如果您的收缩压高于
220 mmHg
臂带开始充气后,按住
START/STOP “加压 / 停止”
按钮,直到血压计充气比预期收缩压高出
30 到
40 mmHg。
注意:
• 血压计不会充气超过 299 mmHg。
• 施压应适可而止,不要过多。
双检测系统
按
START/STOP “加压/停止”按钮,双检测系统指示灯亮
起,装置开始检测设备错误。如果装置准确且运作正常,双检
测系统灯将在测量期间持续亮起。如果检测到系统出错,双检
测
系统灯将闪烁,显示屏上出现 “
Er”。
15
HEM-7211-C1_A_M_CS.fm 16 ページ 2010年2月19日 金曜日 午後8時3分
3. 使用血压计
臂带缠绕指示
臂带缠绕指示这一特征表明臂带是否紧紧缠绕在手臂上。即使显
示 ,血压计仍会进行血压测量。
注意:
由于臂带缠绕有误,此时的测量读数不可靠。请重新缠绕
臂带,务必要正确缠绕臂带,然后再次进行血压测量。显
示 时,臂带紧贴于手臂且缠绕正确,此时的测量读数
准确可靠。
2. 取下臂带。
3. 按 START/STOP “加压 / 停止”按钮关闭血压计。
血压计会自动将测量结果储存在记忆中。
2 分钟后,血压计会自动关闭。
重要事项:
•
最新研究表明以下数值可作为家庭测量时
衡量高血压的指标。
收缩压
舒张压
135 mmHg 以上
85 mmHg 以上
此标准适用于家庭血压测量。
16
HEM-7211-C1_A_M_CS.fm 17 ページ 2010年2月19日 金曜日 午後8時3分
• 您的血压计可侦测非正规心律。非正规心律可能
影响测量结果。非正规心律计算法可自动确定测
量是否有效或需要重复测量。如果测量结果受非
正规心律影响但仍然有效,则显示屏上将显示测
数以及一个非正规心律符号 ( )。如果非正规心
律导致测量无效则不会显示结果。如果在进行测
量后显示非正规心律符号 ( ),请重复测量过
程。如果非正规心律符号 ( ) 频繁显示,请告
知医生。
•
如果在测量过程中移动手臂,则可能显示移动错
误符号 ( )。保持静坐,重新测量。
3. 使用血压计
注意:再次开始血压测量之前请等待
2 - 3 分钟。两次测量之间
的等待可以让动脉恢复到执行血压测量之前的条件。
警告:
依测量结果自我诊断及治疗非常危险。
请遵从医师指导。
17
HEM-7211-C1_A_M_CS.fm 18 ページ 2010年2月19日 金曜日 午後8時3分
3. 使用血压计
3.4
使用记忆读取功能
血压计最多可自动储存
90 组读数。本血压计也可依据最后三
次测量读数 (
10 分钟内)计算平均读数。如果该时段的记忆
中只有两个读数,则平均值将以这两个读数为基准。如果该时
段的记忆中有一个读数,则平均值将以这个读数为基准。
注意:
•
•
如果记忆已满,血压计会删除最早的读数。
查看在未设置日期和时间情况下测量的读数,会显示 “
-:--”,
而不是日期和时间。
查看平均值
1. 按 “记忆”按钮。
注意:如果记忆中没有储存的测量结果,
则显示右侧显示屏。
18
HEM-7211-C1_A_M_CS.fm 19 ページ 2010年2月19日 金曜日 午後8時3分
3. 使用血压计
查看储存在记忆中的读数
1. 显示平均值时,按按钮 。
记忆数位显示一秒钟,然后显
示脉搏速率。最新一组读数的
编号为 “
1”。
交替显示日期 /
时间
2. 按 或 按钮检视记忆中储存的读数。
: 至较早的读数
: 至较近的读数
要删除所有读数
出现记忆符号 ( ) 时,首先按 “记忆”按钮。按住 “记忆”
按钮时,请按下
START/STOP “加压 / 停止”按钮约 2 - 3 秒
钟。
注意:您不能只删除部分读数。
19
HEM-7211-C1_A_M_CS.fm 20 ページ 2010年2月19日 金曜日 午後8時3分
4. 疑难排解和维修
4. 疑难排解和维修
4.1
标志与错误指示符
错误符号 原因
闪烁
亮着
检测到非正规心律。
测量过程中移动身体。
臂带缠绕不正确
电池电量低。
电池耗尽。
更正
取下臂带。等待
2 - 3 分钟后进行
下一次测量。
重复第
3.3 节中的步骤。如果错误
持续出现,请咨询您的医生。
请仔细阅读并重复第
3.3 章节中的
步骤。
正确缠绕臂带。请参阅第
3.1 节。
您应该提前更换电池。
请参阅第
2.1 节。
您应立即更换电池。
请参阅第
2.1 节。
20
HEM-7211-C1_A_M_CS.fm 21 ページ 2010年2月19日 金曜日 午後8時3分
错误符号
4. 疑难排解和维修
臂带漏气。
原因
气塞已断开。
臂带缠绕不正确。
臂带充气过度。
更正
牢固地插入气塞。
请参阅第
3.1 节。
正确缠绕臂带。
请参阅第
3.1 节。
请更换新的臂带。
参阅第
5 章。
重复测量。测量期间,保持静坐并
且不要说话。
请参阅第
3.3 节。
如果重复出现 “
E2”,请手动为臂
带充气,使之比先前测量结果高出
30 到 40 mmHg。
请参阅第
3.3 节。
手动充气臂带时,臂带充气
超过
299 mmHg。
请勿对臂带充气超过
299 mmHg。
请参阅第
3.3 节。
测量过程中移动身体。
衣物妨碍臂带。
重复测量。测量期间,保持静坐并
且不要说话。
请参阅第
3.3 节。
脱下任何妨碍臂带的衣物。
请参阅第
3.1 节。
设备错误。
联络当地
OMRON 零售店或分销
商。
注意: 非正规心律符号 ( ) 可能连同错误消息一并显示。
21
HEM-7211-C1_A_M_CS.fm 22 ページ 2010年2月19日 金曜日 午後8時3分
4. 疑难排解和维修
4.2
疑难排解
无法测量或读数太低
(太高)。
问题
读数过低 (或过高)。
臂带压力没有上升。
臂带排气太快。
按下按钮时无任何反映。
气管没有正确连接到主
机。
原因
臂带缠绕不正确。
测量过程中移动或谈话。
衣物妨碍臂带。
臂带漏气。
臂带松弛。
臂带充气不够。
电池耗尽。
电池安装不正确。
更正
正确缠绕臂带。
请参阅第
3.1 节。
测量期间,保持静坐并且
不要说话。
请参阅第
3.3 节。
脱下任何妨碍臂带的衣
物。
请参阅第
3.1 节。
确保气管连接牢固。
请参阅第
3.1 节。
更换新的臂带。
参阅第
5 章。
正确缠绕臂带,使其牢固
地缠绕于手臂。
请参阅第
3.1 节。
对臂带充气,使之比先前
测量结果高出
30 到
40 mmHg。
请参阅第
3.3 节。
用新的电池更换旧电池。
请参阅第
2.1 节。
依正确的 (
+/-) 极性安装
电池。
请参阅第
2.1 节。
22
HEM-7211-C1_A_M_CS.fm 23 ページ 2010年2月19日 金曜日 午後8時3分
问题
其它问题。
4. 疑难排解和维修
原因 更正
•
按下
START/STOP“加压 / 停止”按钮,重复测量。
•
如果问题继续存在,尝试用新电池更换旧电池。
如果这依旧不能解决问题, 请联络您当地的
OMRON
授权分销商或零售商
23
HEM-7211-C1_A_M_CS.fm 24 ページ 2010年2月19日 金曜日 午後8時3分
4. 疑难排解和维修
4.3
维修
要保护血压计免遭损坏,请遵守以下内容:
•
请勿将主机和臂带置于高温、高湿、湿润或者阳光直射的环境
中。
•
•
•
•
•
•
•
请勿用力折叠臂带或气管。
请勿拆卸主机。
避免血压计遭受撞击 (例如,将血压计跌落至地面)。
请勿使用挥发性液体清洁主机。
请勿清洗臂带或将其浸入水中。
请勿使用汽油、稀释剂或类似的溶剂清洁臂带。
请勿自行维修。如果出现故障,请联络您当地的
OMRON 授
权分销商或零售商。
24
•
清洁血压计应使用柔软干燥的抹布。
• 使用柔软的湿布和肥皂清洁臂带。
HEM-7211-C1_A_M_CS.fm 25 ページ 2010年2月19日 金曜日 午後8時3分
4. 疑难排解和维修
校准与服务
•
• 本血压计的精度经过精心检测,且使用寿命长。
通常建议每隔两年对血压计进行检查,以确保其功能正常、准
确无误。请联络您当地的
OMRON 授权分销商或零售商。
4.4
储存
血压计不使用时要放入储存盒中。
1. 从充气孔拔下气塞。
2. 轻轻将气管摺叠进臂带。
注意:请勿过度弯曲气管。
3. 将臂带和主机放入储存盒中。
请勿在以下情况储存血压计 :
•
•
如果血压计潮湿。
高温、潮湿、阳光直射、存在灰尘
或腐蚀性气体的场所。
• 易晃动、震动或形成角度的位置。
保养和维修
25
HEM-7211-C1_A_M_CS.fm 26 ページ 2010年2月19日 金曜日 午後8時3分
5. 选件
5. 选件
中型臂带
手臂周长
22 - 32 厘米
大型臂带
手臂周长
32 - 42 厘米
小型臂带
手臂周长
17 - 22 厘米
交流适配器
S
使用交流适配器选件
1. 将交流适配器插头插入主机背面
的交流适配器插孔。
2. 将交流适配器插入电源插座。
要断开交流适配器,请首先从电源插座上拔出交流适配器,然
后将交流适配器从主机取下。
26
HEM-7211-C1_A_M_CS_1.fm Page 27 Wednesday, June 2, 2010 9:54 AM
6. 技术资料
6. 技术资料
产品名称
型号
显示
测量方式
测量范围
准确度
充气
排气
记忆
额定
电源
自动血压计
HEM-7211 (HEM-7211-C1(VG))
LCD 数位显示屏
动脉搏动描记法
血压:
0 至 299 mmHg
脉搏:
40 至 180/ 分钟
血压:
±3 mmHg
显示读数的
±5%
电动气泵模糊逻辑控制
自动减压阀
90 组测量值及其日期和时间
DC6V 4W
4 节 “AA”电池 1.5V 或交流适配器 (交流适配器 -
9515336-9
( 选件 ), 电流
AC100-240V 50/60Hz
S
0.12A
)
全新碱性电池大约可测量
1500 次
电池寿命
应用的部件:
= B 型
电击保护
操作温度 / 湿度
存放温度 / 湿度 / 气压
主机重量
臂带重量
外形尺寸
臂带尺寸
臂带材料
附件
内部供电的
ME 设备 (仅使用电池时)
=
II 级 ME 设备 (可选交流适配器)
+10°C 至 +40°C / 最高:相对湿度 30% 至 85%
-20°C 至 +60°C / 最高:10% 至 95% RH/ 700-1060 hPa
约
380 克 (不含电池)
约
120 克
约
123 毫米 (宽) x 85 毫米 (高) x 158 毫米 (长)
约
146 毫米 x 446 毫米 (中型臂带:手臂周长 22 - 32 厘米)
尼龙和聚酯
主机、臂带、使用说明书、储存盒、电池
注意:
• 技术资料如有更改,恕不提前通知。
• 本产品和废旧电池的处理应遵循国内对电子产品的处理规定。
这
OMRON 产品是在 OMRON HEALTHCARE Co., Ltd.,日本,严格的质量体系
下生产的。
OMRON 血压计的核心电子原部件,所谓的压力传感器,是在日本生产
组装。
27
HEM-7221-C1_C_M.book Page 29 Monday, November 22, 2010 1:19 PM
Asia Pacific HQ
Production Facility
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp
THE NETHERLANDS
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08
Alexandra Technopark, Singapore 119967 www.omron-healthcare.com.sg
OMRON HEALTHCARE MANUFACTURING VIETNAM
Binh Duong Province, VIETNAM
Co.,Ltd.
5334767-8A
Download
Advertisement
Key features
IntelliSense technology
Dual Check System
Cuff Wrapping Guide
Memory function
Automatic inflation and deflation
Compact and easy to use
Accurate and reliable measurements
90 readings memory
Frequently asked questions
IntelliSense technology is an advanced feature that allows for comfortable controlled inflation without the need of pressure presetting or re-inflation.
The Dual Check System is a feature that helps to ensure the accuracy of the blood pressure measurement by detecting device errors.
The Cuff Wrapping Guide is a unique feature that indicates if the cuff is not wrapped tightly enough around the arm. Even when the icon is displayed, a blood pressure reading will be taken. This reading is NOT reliable due to the incorrect wrapping of the cuff. Please wrap the cuff again, taking care to wrap it correctly and take the measurement again. When the icon is displayed, the cuff is correctly wrapped tightly enough on the arm and the reading is accurate and reliable.
The HEM-7211 can store up to 90 readings with date and time.
To replace the batteries, remove the battery cover, insert four “AA” batteries as indicated in the battery compartment and then replace the battery cover.