advertisement
![Metrohm 831 KF Coulometer and 860 KF Thermoprep Owner's Manual | Manualzz Metrohm 831 KF Coulometer and 860 KF Thermoprep Owner's Manual | Manualzz](http://s2.manualzz.com/store/data/062106370_1-1a68c3caa63b1d8701e425bb7d583bf0-360x466.png)
860 KF Thermoprep
Manual
8.860.8003EN / 2020-03-23
Metrohm AG
CH-9100 Herisau
Switzerland
Phone +41 71 353 85 85
Fax +41 71 353 89 01 [email protected]
www.metrohm.com
860 KF Thermoprep
Manual
8.860.8003EN / 2020-03-23
Technical Communication
Metrohm AG
CH-9100 Herisau [email protected]
This documentation is protected by copyright. All rights reserved.
This documentation has been prepared with great care. However, errors can never be entirely ruled out. Please send comments regarding possible errors to the address above.
■■■■■■■■■■■■■■■■■■■■■■
Table of contents
Table of contents
1
Instrument description ......................................................... 1
Instrument components ........................................................... 1
Intended use ........................................................................... 2
About the documentation ................................................... 2
Symbols and conventions ........................................................ 2
Safety instructions ................................................................ 3
General notes on safety ........................................................... 3
Electrical safety ........................................................................ 3
Personnel safety ....................................................................... 4
Flammable samples and solvents ............................................. 5
Recycling and disposal ............................................................. 5
6
8
Setting up the instrument .................................................... 8
Packaging ................................................................................ 8
Checks .................................................................................... 8
Location .................................................................................. 8
Connecting the instrument to the power grid ................... 8
Mounting the support rod ................................................... 9
Mounting the sample holder ............................................. 12
Mounting the needles ........................................................ 13
Mounting the heating tubing ............................................ 15
Mounting the dust filter .................................................... 16
Assembling drying flasks ................................................... 17
Assembling the air/nitrogen connector ............................ 19
3.10 Inserting the heating tube into the KF titration cell ....... 19
Switching the instrument on and off ................................ 22
The keypad .......................................................................... 22
The display .......................................................................... 23
Display elements .................................................................... 23
Status display ......................................................................... 23
Error messages ...................................................................... 24
860 KF Thermoprep
■■■■■■■■ III
Table of contents
■■■■■■■■■■■■■■■■■■■■■■
Modifying numerical values and settings ......................... 24
Selecting from a selection list ................................................. 25
Modifying numerical values ................................................... 25
Parameters .......................................................................... 25
Configuration ...................................................................... 27
5 Performing a determination 28
Conditioning the system .................................................... 28
Preparing the KF titration cell ................................................. 28
Preparing the 860 KF Thermoprep ......................................... 29
Determining the blank value ............................................. 32
Determining samples ......................................................... 33
6 Operation and maintenance 35
General ................................................................................ 35
Care ...................................................................................... 35
Problems and their solutions ............................................. 36
Practical notes .................................................................... 37
Literature ............................................................................. 38
Oven ..................................................................................... 39
Gas flow .............................................................................. 39
Outlet heater ....................................................................... 39
Ambient conditions ............................................................ 40
Energy supply ...................................................................... 40
Reference conditions .......................................................... 40
Dimensions .......................................................................... 41
IV ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
Table of figures
Table of figures
Front 860 KF Thermoprep ................................................................. 6
Rear 860 KF Thermoprep ................................................................... 7
Loosening the adapter ....................................................................... 9
Mounting the support rod ............................................................... 10
Mounting the tubing ....................................................................... 11
Attaching the adapter ..................................................................... 11
Mounted support rod ...................................................................... 12
Mounting the sample holder ........................................................... 12
Mounting the needles ..................................................................... 14
Mounting the heating tubing (rear view) ......................................... 15
Connecting the heating tubing ........................................................ 16
Mounting the dust filter .................................................................. 16
Preparing the drying flasks ............................................................... 17
Mounting the tubings ..................................................................... 18
External gas connector .................................................................... 19
Coulometric KF titration cell ............................................................ 20
Volumetric KF titration cell ............................................................... 21
Power switch .................................................................................. 22
Keypad of the operating unit ........................................................... 22
Main dialog ..................................................................................... 23
Insert the sample vessel into the oven ............................................. 30
Move the guide head down ............................................................ 30
Insert the needle ............................................................................. 31
860 KF Thermoprep
■■■■■■■■ V
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction
1 Introduction
1.1 Instrument description
The 860 KF Thermoprep is used wherever the heating of a sample and/or the thermal expulsion of moisture in solids or liquids is required. In combination with a coulometric or volumetric KF titrator, the 860 KF Thermoprep is the ideal analysis system for water content determination in samples that contain disruptive components or from which moisture can be removed only with difficulty.
One of its decisive advantages is reducing sample preparation to a minimum. Thanks to the use of hermetically sealed sample vessels ("headspace vials"), samples can be filled directly on-site. The PTFE-coated septa guarantee a constant, unaltered water content, even after prolonged holding times.
The sample heated in the oven module releases its moisture in the form of water vapor, which is conveyed into a measuring cell with the aid of a gas flow. An air pump is installed to generate the gas flow. An inlet valve is available for nitrogen or other inert gases. The moisture can be determined in the measuring cell either coulometrically or volumetrically using
Karl Fischer titration.
1.1.1 Instrument components
The 860 KF Thermoprep has the following components:
■
■
■
■
■
■
■
Oven
Oven module made of aluminum with software-operated temperature control for heating the sample vessel.
Fan
Propeller fan for cooling the oven module.
Inlet valve
Valve for switching over the source of the gas flow.
Air pump
Pump for generating the gas flow.
Outlet heater
Heating tubing for preventing the condensation of moisture.
Support rods
Guidance device with needle adapter and tubing for the gas flow.
Operating unit
Monochrome LCD display and keyboard.
860 KF Thermoprep
■■■■■■■■ 1
1.2 About the documentation
■■■■■■■■■■■■■■■■■■■■■■
1.1.2 Intended use
The 860 KF Thermoprep is designed for usage as an auxiliary device for sample preparation in analytical laboratories. Its main area of application is water content determination using Karl Fischer titration (coulometric or volumetric). The 860 KF Thermoprep enables the application of thermal gas extraction technology.
This instrument is suitable for processing chemicals and flammable samples. Therefore, the use of the 860 KF Thermoprep requires the user to have basic knowledge and experience in handling toxic and caustic substances. Knowledge regarding the application of fire prevention measures prescribed for laboratories is also mandatory.
1.2 About the documentation
CAUTION
Please read through this documentation carefully before putting the instrument into operation. The documentation contains information and warnings which the user must follow in order to ensure safe operation of the instrument.
1.2.1 Symbols and conventions
The following symbols and formatting may appear in this documentation:
Method
File ▶ New
[Next]
Cross-reference to figure legend
The first number refers to the figure number, the second to the instrument part in the figure.
Instruction step
Carry out these steps in the sequence shown.
Dialog text , parameter in the software
Menu or menu item
Button or key
WARNING
This symbol draws attention to a possible life-threatening hazard or risk of injury.
WARNING
This symbol draws attention to a possible hazard due to electrical current.
2 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
1.3 Safety instructions
1.3.1 General notes on safety
WARNING
1 Introduction
WARNING
This symbol draws attention to a possible hazard due to heat or hot instrument parts.
WARNING
This symbol draws attention to a possible biological hazard.
CAUTION
This symbol draws attention to possible damage to instruments or instrument parts.
NOTE
This symbol highlights additional information and tips.
Operate this instrument only according to the information contained in this documentation.
This instrument left the factory in a flawless state in terms of technical safety. To maintain this state and ensure non-hazardous operation of the instrument, the following instructions must be observed carefully.
1.3.2 Electrical safety
The electrical safety when working with the instrument is ensured as part of the international standard IEC 61010.
WARNING
Only personnel qualified by Metrohm are authorized to carry out service work on electronic components.
860 KF Thermoprep
■■■■■■■■ 3
1.3 Safety instructions
■■■■■■■■■■■■■■■■■■■■■■
WARNING
Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched.
There are no parts inside the housing which can be serviced or replaced by the user.
Supply voltage
WARNING
An incorrect supply voltage can damage the instrument.
Only operate this instrument with a supply voltage specified for it (see rear panel of the instrument).
Protection against electrostatic charges
WARNING
Electronic components are sensitive to electrostatic charges and can be destroyed by discharges.
Do not fail to pull the power cord out of the power socket before you set up or disconnect electrical plug connections at the rear of the instrument.
1.3.3 Personnel safety
WARNING
Wear protective glasses and working clothes suitable for laboratory work while operating the 860 KF Thermoprep.
WARNING
The oven and the sample vessels can reach temperatures of up to
250 °C ! Avoid direct skin contact. Use the provided septum crimping tongs or another suitable holding device to insert the sample vessels into or remove them from the oven. Wear heat-insulating gloves, if necessary.
4 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction
WARNING
The needles pose a considerable risk of injury . When inserting the needle into the sample vessel, hold the handle of the needle adapter with both hands. Ensure that no other person reaches into the danger zone during this procedure.
1.3.4 Flammable samples and solvents
WARNING
■
■
■
■
■
All relevant safety measures are to be observed when working with flammable samples or when adding solvents.
Use nitrogen or another inert gas for conveying the moisture.
Set up the instrument in a well-ventilated location (e.g., fume cupboard).
Keep all sources of flame far from the workplace.
Clean up spilled liquids and solids immediately.
Follow the safety instructions of the chemical manufacturer.
1.3.5 Recycling and disposal
This product is covered by European Directive 2012/19/EU, WEEE – Waste
Electrical and Electronic Equipment.
The correct disposal of your old instrument will help to prevent negative effects on the environment and public health.
More details about the disposal of your old instrument can be obtained from your local authorities, from waste disposal companies or from your local dealer.
860 KF Thermoprep
■■■■■■■■ 5
■■■■■■■■■■■■■■■■■■■■■■
2 Overview of the instrument
1
2
3
8
4
5
6
7
1 Guide head
Figure 1 Front 860 KF Thermoprep
2 Needle holder (6.2049.040)
With needle adapter.
4 Guidance rod 3 Spring sleeve
With vessel stripper.
5 Sample vessel (6.2419.007)
With 6 ml content.
7 Oven module
With sheet metal housing.
9 Operating unit
With display and keyboard.
11 Drying flask holder
6 Sample holder (6.2063.010)
8 Fixing lever
10 Drying flasks (6.1608.050)
With 6.1602.145 drying flask insert.
12 Gas inlet
With M6 thread.
13 Gas outlet
With M6 thread.
9
10
11
12
13
6 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
2 Overview of the instrument
8
1
2
3
4 5 6
Figure 2 Rear 860 KF Thermoprep
2 Mains switch 1 Air pump inlet
With 6.2724.010 dust filter.
3 Mains connection socket
5 Outlet heating connection
For the heating tube.
7 Fan
For cooling the oven module.
4 Type plate
Contains specifications concerning mains voltage and serial number.
6 Air/nitrogen connector
With M6 interior thread. Inlet for external gassing.
8 Cable holder (6.2060.010)
7
860 KF Thermoprep
■■■■■■■■ 7
3.1 Setting up the instrument
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
3.1 Setting up the instrument
3.1.1 Packaging
The instrument is supplied in protective packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument.
3.1.2 Checks
Immediately after receipt, check whether the shipment has arrived complete and without damage by comparing it with the delivery note.
3.1.3 Location
The instrument has been developed for operation indoors and may not be used in explosive environments.
Place the instrument in a location of the laboratory which is suitable for operation and free of vibrations and which provides protection against corrosive atmosphere and contamination by chemicals.
The instrument should be protected against excessive temperature fluctuations and direct sunlight.
3.2 Connecting the instrument to the power grid
WARNING
Electric shock from electrical potential
Risk of injury by touching live components or through moisture on live parts.
■
■
■
■
Never open the housing of the instrument while the power cord is still connected.
Protect live parts (e.g. power supply unit, power cord, connection sockets) against moisture.
Unplug the power plug immediately if you suspect that moisture has gotten inside the instrument.
Only personnel who have been issued Metrohm qualifications may perform service and repair work on electrical and electronic parts.
8 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
Accessories
Connecting the power cord
■
■
■
■
■
Power cord with the following specifications:
Length: max. 2 m
Number of cores: 3, with protective conductor
Instrument plug: IEC 60320 type C13
Conductor cross-section 3x min. 0.75 mm 2 / 18 AWG
Power plug:
– according to customer requirement (6.2122.XX0)
– min. 10 A
NOTICE
Do not use a not permitted power cord!
1 Plugging in the power cord
■
■
Plug the power cord into the instrument's power socket.
Connect the power cord to the power grid.
3.3 Mounting the support rod
The support rod with the guide head is enclosed separately with the 860
KF Thermoprep and must first be mounted.
Proceed as follows:
1 Move the guide head of the support rod a bit downwards by pressing the locking lever.
3
1 2
860 KF Thermoprep
Figure 3 Loosening the adapter
2 Loosen the adapter on the head of the support rod with a hex key.
■■■■■■■■ 9
3.3 Mounting the support rod
■■■■■■■■■■■■■■■■■■■■■■
3 Removing the adapter.
4 Insert the tubing jutting out of the housing from below into the support rod.
NOTICE
This tubing conveys the gas from the flow controller upward through the support rod to the guide head. From there, the gas is guided to the double hollow needle (see next chapter).
5
4
Figure 4 Mounting the support rod
5 Insert the support rod into the opening intended for this purpose from above and push it downwards until it hits the stop. Observe the orientation of the support rod.
6 Pull out the tubing inside the support rod as far as possible. You might want to use a piece of sand paper to be able to better grasp the tubing.
10 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
7
6
3 Installation
Figure 5 Mounting the tubing
7 Fasten the tubing on the olive of the 4.860.4260 adapter.
8 Pull out the support rod as far as possible so that the adapter rests closely on the support rod.
8
9
10
Figure 6 Attaching the adapter
9 Attach the adapter to the support rod. Fasten the screw into the support rod with the aid of a hex key.
10 Carefully push the support rod downwards.
860 KF Thermoprep
■■■■■■■■ 11
3.4 Mounting the sample holder
■■■■■■■■■■■■■■■■■■■■■■
11
Figure 7 Mounted support rod
11 Fix the support rod with the V.0434.008 hex screw following the illustration. The screw and the corresponding hex key are enclosed separately with the instrument.
3.4 Mounting the sample holder
6.2063.010
6.2627.000
Figure 8 Mounting the sample holder
NOTICE
The sample holder may not be inserted or removed unless it has cooled down.
12 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
The dimensions of the supplied sample holder are optimized for the usage of 6.2419.007 6-mL sample vials by Metrohm. This ensures the best possible transfer of heat between oven and sample.
1 Place the 6.2627.000 oven insert into the oven from above.
2 Place the 6.2063.010 sample holder into the oven from above.
■
■
If vessels with other dimensions are to be used, then individually modified sample holders can be ordered. The precise vessel dimensions (including tolerances) are required. Non-standard sample vessels may not exhibit dimensions outside of the following limit values:
Diameter 10.0 - 32.0 mm
Immersion depth 20.0 - 45.0 mm
3.5 Mounting the needles
The length of the needle holder defines how deep the needle for the gas inlet enters the sample vessel. The 6.2049.040 needle holder with a length of 58 mm ensures that the needle penetrates the liquid or powdery sample. The carrier gas flows through the sample and efficiently removes the contained moisture.
If the heated sample may block the needle, use the 6.2049.050 needle holder with a length of 73 mm. In this case, the needle for the gas inlet enters the sample vessel slightly more than the outlet needle and it does not have any contact with the sample. The 6.2049.050 needle holder can be ordered at Metrohm if required.
Needle holders with the dimensions required for situations calling for special sample vessels can be supplied by Metrohm upon request.
860 KF Thermoprep
■■■■■■■■ 13
3.5 Mounting the needles
■■■■■■■■■■■■■■■■■■■■■■
6.1805.100
6.2816.070
3
4 4
6.2049.040
1
6.2816.080
2
Figure 9 Mounting the needles
Mount the needles as follows:
1 Screw the 6.2049.040 needle holder into the distributor on the guide head.
2 Screw the 6.2816.080 outlet needle onto the Luer connector of the needle holder.
3 Carefully guide the 6.2816.070 needle for gas inlet into the opening of the distributor from above and allow it to drop down.
NOTICE
Ensure that the white PTFE seal sits firmly on the needle.
4 Screw the 6.1805.100 FEP tubing onto the opening of the distributor by hand. Screw the other end of the tubing onto the opening on the upper side of the support rod.
14 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
3.6 Mounting the heating tubing
3 Installation
6.2060.010
3
2
1 6.1830.030
860 KF Thermoprep
Figure 10 Mounting the heating tubing (rear view)
Proceed as follows:
1 Screw the M6 connector of the 6.1830.030 heating tubing into the lateral opening of the distributor on the guide head.
2 Use the accompanying knurled screw to screw the 6.2060.010
cable holder tightly onto the guide head from below.
3 Guide the heating tubing cable through the eyelet of the cable holder.
NOTICE
The cable holder prevents the cable of the heating tubing from coming into contact with parts of the housing that are hot.
4 Connect the heating tubing cable to the Outlet heater connector on the rear of the instrument.
■■■■■■■■ 15
3.7 Mounting the dust filter
■■■■■■■■■■■■■■■■■■■■■■
4
Figure 11 Connecting the heating tubing
Rotate the plug in such a way that the three contact pins correspond to the sequence of the corresponding socket openings. Press the plug towards the socket and rotate the front knurled screw clockwise.
NOTICE
The heating jacket of the heating tubing is heated up to approx.
40 to 50 °C as soon as the instrument is switched on. This prevents condensation of moisture in the tubing when it is removed from the sample and transferred to a KF titration cell with a carrier gas.
3.7 Mounting the dust filter
The installed air pump must be protected from dust. For this reason, a
6.2724.010 dust filter must be placed on the air inlet ( Inlet filter ) on the right-hand side of the housing..
6.2724.010
Figure 12 Mounting the dust filter
The dust filter should be replaced once a year.
16 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
3.8 Assembling drying flasks
Two drying flasks with desiccant are integrated into the gas flow to dry the supplied gas. Dust (e.g. from the desiccant) must be prevented from finding its way into the sample vessel.
6.1602.145
6.1821.040
2
3
1
6.2811.000
6.1608.050
Figure 13 Preparing the drying flasks
Prepare the two drying flasks as follows:
1 Fill both 6.1608.050 drying flasks with 6.2811.000 molecular sieve .
2 Screw one 6.1821.040 filter tube into each of the 6.1602.145
drying flask caps from below. Tighten the filter tubes by hand.
3 Screw the two drying flask caps with the filter tubes onto the drying flasks. Tighten the caps by hand.
NOTICE
If drying flask caps or filter tubes are not tightened well, the gas flow may not be precise and regular. The error message "Flow rate error" usually occurs when screw connections are not tight.
860 KF Thermoprep
■■■■■■■■ 17
3.8 Assembling drying flasks
6.1805.080
2
6.1805.010
4
6.1805.080
3
■■■■■■■■■■■■■■■■■■■■■■
1
1
3
2
Figure 14 Mounting the tubings
Mount the FEP tubings as follows:
1 Insert the two drying flasks that have been prepared into the holders, see illustration.
2 Screw one 6.1805.080 FEP tubing (25 cm length) to the gas outlet
(at the front, labeled 1 ). Screw the other tubing end on the front drying flask to the M6 connector without point marking (see left-hand arrow).
3 Screw the second 6.1805.080 FEP tubing (25 cm length) to the gas inlet (at the rear, labeled 2 ). Screw the other tubing end on the rear drying flask to the M6 connector with point marking (see right-hand arrow).
4 Screw the 6.1805.010 FEP tubing (13 cm length) tightly onto the remaining M6 connectors of the drying flasks.
The Drying flask marking on the right-hand side of the instrument displays the diagram for the tubing.
NOTICE
Tighten the screw connections well by hand.
18 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
3.9 Assembling the air/nitrogen connector
If compressed air, nitrogen or another gas is to be used for transferring the expelled moisture, then a separate connector is available at the rear of the instrument.
A tube with M6 thread can be connected directly to the Air/N
2
in connector. Enclosed with the instrument is the 6.1808.040 M6/M8 tubing adapter for a tubing with an M8 thread. The 6.1808.050 M8/tubing olive can additionally be put in place in order to connect a simple tubing.
6.1808.040
6.1808.050
Figure 15 External gas connector
WARNING
If gas is fed from a pressure line or a pressure vessel, a pressure reduction valve must be activated. The gas pressure must not exceed 0.3 bar of overpressure.
3.10 Inserting the heating tube into the KF titration cell
The 860 KF Thermoprep can be used in combination with a KF coulometer or a volumetric KF titrator. The mounting of the tip of the heating tube is accomplished in different ways for the respective KF titration cells.
Coulometric KF titration cell
■
■
Remove the protective cover from the tip of the heating tube and the
E.3010.032 O-ring.
Disassemble the accompanying 6.1446.170 heating tube stopper into three parts.
860 KF Thermoprep
■■■■■■■■ 19
3.10 Inserting the heating tube into the KF titration cell
6.1830.030
2
1
3
■■■■■■■■■■■■■■■■■■■■■■
6.1446.170
20 ■■■■■■■■
Figure 16 Coulometric KF titration cell
Proceed as follows:
1 Guide the upper part of the heating tube stopper over the tip of the
6.1830.030 heating tube as shown in the illustration.
2 Guide the O-ring of the heating tube stopper over the heating tube.
3 Guide the lower part of the heating tube stopper over the heating tube and screw the three parts together (not too tightly).
4 Insert the stopper with the tip of the heating tube into one of the two SGJ openings of the KF titration cell
5 Shift the tip vertically in such a way that the outlet opening of the tubing is immersed as deeply as possible. The tip must not however be permitted to get in the way of the stirring bar in the KF titration cell. Afterwards give the heating tube stopper its final tightening.
■
■
Volumetric KF titration cell
Remove the protective cover from the tip of the heating tube and the
E.3010.032 O-ring. The latter you will still need.
A 6.2730.020 septum stopper is enclosed with every KF titrator made by Metrohm. Disassemble this septum stopper into three parts and remove the septum. It will not be required.
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
6.1830.030
E.3010.032
1
2
3
6.2730.020
3 Installation
Figure 17 Volumetric KF titration cell
Proceed as follows:
1 Guide the upper part of the septum stopper over the tip of the
6.1830.030 heating tube as shown in the illustration.
2 Guide the E.3010.032
O-ring over the heating tube.
This O-ring is also part of the 6.1244.040 set of seals that is enclosed with each KF titrator. It cannot be reordered individually.
3 Guide the lower part of the septum stopper over the heating tube and screw the three parts together (not too tightly).
4 Insert the stopper with the tip of the heating tube into the front opening of the KF titration cell
5 Shift the height of the tip of the heating tube vertically in such a way that the outlet opening of the tubing is immersed as deeply as possible. The tip must not however be permitted to get in the way of the stirring bar in the KF titration cell. Afterwards give the septum stopper its final tightening.
860 KF Thermoprep
■■■■■■■■ 21
4.1 Switching the instrument on and off
■■■■■■■■■■■■■■■■■■■■■■
4 Operation
4.1 Switching the instrument on and off
The on/off switch is located on the rear of the instrument.
Figure 18 Power switch
The oven is switched on automatically after the instrument has been switched on and it is heated to the temperature that was set last. At the same time, the outlet heating of the heating tubing is also switched on.
4.2 The keypad
22 ■■■■■■■■
Figure 19 Keypad of the operating unit
Key functions in the main dialog
[FLOW] Switches the gas flow on or off.
[TEMP]
[QUIT]
[OK]
[ ◀ ]
Switches between the two target temperatures.
No function.
Opens parameter menu.
Reduces the gas flow by 1 mL/min during operation.
Press the key shortly.
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
[ ▶ ]
[ ▲ ]
[ ▼ ]
4.3 The display
4.3.1 Display elements
4 Operation
Increases the gas flow by 1 mL/min during operation.
Press the key shortly.
Reduces the brightness of the display.
Increases the brightness of the display.
1
2
3
4
5
Figure 20 Main dialog
1 Temperature display
Shows the measured oven temperature.
3 Gas flow display
Shows the set gas supply and the flow rate.
2 Icon of the status display
4 Target temperature
Shows the set specification (temp. 1 or temp. 2).
5 Status display
Shows status and error messages.
4.3.2 Status display
Table 1 Icons of the status display
Icon Meaning
The oven is heated.
The oven is cooled.
The predefined temperature has been reached.
An error has occurred. The error type is shown in the status display.
860 KF Thermoprep
■■■■■■■■ 23
4.4 Modifying numerical values and settings
■■■■■■■■■■■■■■■■■■■■■■
4.3.3 Error messages
Table 2 Error messages
Error message Cause
Flow rate error The tubing system is leaking or the flow rate is set too high.
Max. temp.
reached
Adj. data missing
The pressure is too low or too high during external gas supply.
The flow sensor is defective.
The oven temperature is higher than
265 °C. The temperature control might be defective.
The adjustment data of the instrument is not correct.
Remedy
Tighten all tubing connections and screws of the drying flasks.
Check the function of the pump.
Regulate the pressure of the external gas to 0.1 -
0.3 bar.
Contact a service engineer.
Switch off the instrument immediately and let the oven cool down.
If the error reoccurs again when switching the instrument on, contact a service engineer.
The instrument must be adjusted again. Contact a service engineer.
4.4 Modifying numerical values and settings
The settings of the instrument can be modified in two ways:
■ Selection of predefined values from a list. Parameters with selection lists are marked with a colon ( : ).
Modifying numerical values.
■
The various parameters can be selected using the [ ▼ ] and [ ▲ ] arrow keys.
The current line will be displayed reversed.
24 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
4 Operation
4.4.1 Selecting from a selection list
You can modify parameters with selection lists (e.g. Flow source: ) using the [ ▶ ] or [ ◀ ] arrow keys. The next value in the selection list is usually displayed.
Apply the modified value with the [OK] key.
4.4.2 Modifying numerical values
Modify parameters with numerical values using the arrow keys:
1 Use the [ ▶ ] or [ ◀ ] arrow keys to select a digit of the numerical value.
The selected digit is displayed in black on white. The colors for the rest of the line remain reversed.
2 Modify the digit using the [ ▲ ] or [ ▼ ] arrow keys.
3 Select the next digit and modify it, etc.
4 Apply the modified value with the [OK] key.
Press [QUIT] if the previous value is to be restored.
4.5 Parameters
[OK]
The instrument parameters can be set in a menu which can be opened with the [OK] key. Quit the parameter menu with [QUIT] .
NOTICE
Modified values are applied by pressing [OK] . A modification is discarded with [QUIT] .
Temp. 1
Default value for temperature 1.
Input range
Default value
50 - 250 °C (Increment: 1)
100 °C
Temp. 2
Default value for temperature 2.
860 KF Thermoprep
■■■■■■■■ 25
4.5 Parameters
Gas flow
Gas type
Meas. factor
Flow source
50 - 250 °C (Increment: 1)
100 °C
■■■■■■■■■■■■■■■■■■■■■■
Input range
Default value
Default value for the gas flow.
Input range
Default value
10 - 150 mL/min (Increment: 1)
50 mL/min
Selection of the gas.
Selection
Default value air | nitrogen | other air other
A measuring factor must be specified for the gas used when selecting other . The measuring factor is used for the correct measurement of the flow rate.
Measuring factor for measuring the flow rate of the gas. This setting is only possible if other was selected as gas. For air and nitrogen, the corresponding measuring factor is inserted automatically.
Only if 'gas type' = 'other'.
Input range
Default value
0.001 - 2.000 (Increment: 0.001)
1.000
Selection of the gas supply.
Selection
Default value pump pump
| valve pump
Built-in air pump.
valve
Inlet valve for inert gas.
26 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
4 Operation
4.6 Configuration
[TEMP]
The basic settings of the instrument can be configured in the setup menu.
You can open the menu by pressing the [TEMP] key while switching the instrument on. Press [QUIT] to close the menu.
NOTICE
The setup menu can be opened even if the key lock is on.
Lock keys
Key lock. The [OK] , [ ◀ ] and [ ▶ ] keys can be locked. With this, you can prevent that changes are made in the parameter menu or that the gas flow is changed during a determination. If this function is enabled ( on ), you can only switch between temperature 1 and temperature 2 and switch the gas flow on or off. The brightness of the display can still be adjusted.
Selection
Default value off | on off
Temp.corr
Temperature correction. The default value for the temperature control of the oven can be influenced by this correction value. In this way, you can adjust a temperature difference between the oven module and the sample in the vessel if required.
Input range
Default value
-10 - 10 °C (Increment: 1)
0 °C
The determination of the necessary temperature correction can be accomplished with a special oven insert and should be performed by a service engineer. Contact your responsible Metrohm supplier.
860 KF Thermoprep
■■■■■■■■ 27
5.1 Conditioning the system
■■■■■■■■■■■■■■■■■■■■■■
5 Performing a determination
Moisture determinations according to the heating method require the conditioning of the entire tubing system and of the KF titration cell before a determination can be carried out. Because the sample vials and the septum seals may contain small amounts of moisture, three to five blank value determinations must be carried out with sealed, empty vessels.
Solid or liquid samples are weighed out into sample vials, which are then sealed. We recommend the use of 6.2419.007 sample vials with
6.1448.057 septum seals. The silicone septum of the aluminum cap resists temperatures of up to 250 °C and has proven itself to be an optimum solution.
5.1 Conditioning the system
The entire system must be conditioned prior to use, i. e. it must be absolutely water-free. The oven must be heated to the desired working temperature for this purpose. The needle must be inserted into a sealed sample vessel (conditioning vessel) in order to flush the entire tubing system with the carrier gas. The gas supply is thus connected with the heating tube.
5.1.1 Preparing the KF titration cell
Filling the KF titration cell
The KF titration cell must remain filled with conditioned working medium when not in use. The working medium must be replaced from time to time, even though the sample is not transferred into the working medium during the heating-out process.
Prepare the KF titration cell as follows:
1 Fill the KF titration cell with working medium. We recommend approximately 150 mL of working medium when using a Coulometer cell and approximately 35 mL with a volumetric KF titration cell. The shaft of the heating tube should, however, not be immersed.
2 Align the tip of the heating tube against the vessel wall.
3 Switch on the stirrer. Set the stirring rate to high.
28 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
5 Performing a determination
It is advantageous to have air bubbles stirred into the working medium and forcefully broken up in the process. The air in the KF titration cell must also be dried.
Conditioning the working medium
1 Load a conditioning method on the KF titrator and begin the conditioning by pressing the [Start] key.
You will find the necessary settings in the KF titrator manual.
NOTICE
It is imperative that an extraction time of at least 300 seconds be set for a determination using the heating method.
5.1.2 Preparing the 860 KF Thermoprep
Switch on the 860 KF Thermoprep and select the temperature
The necessary settings on the 860 KF Thermoprep:
1 Switch on the instrument.
2 Select one of the preset temperatures using the [TEMP] key.
If you would like to change Temp. 1 or Temp. 2 , press [OK] .
Inserting the conditioning vessel
1 Seal an empty sample vessel tightly with a septum seal. To do this, use the crimping tongs.
2 Insert the sample vessel into the oven.
NOTICE
Caution!
The oven is hot. Use the crimping tongs to insert the sample vessel.
860 KF Thermoprep
■■■■■■■■ 29
5.1 Conditioning the system
■■■■■■■■■■■■■■■■■■■■■■
Figure 21 Insert the sample vessel into the oven
Inserting the needle
1 Grip the guide head with both hands and press the fixing lever.
Figure 22 Move the guide head down
2 Move the guide head down and insert the needle into the sample vessel.
30 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
5 Performing a determination
860 KF Thermoprep
Figure 23 Insert the needle
Switching on the gas flow
1 Press the [OK] key and adjust the flow rate ( Gas flow ), e. g.
50 mL/min. Use the arrow keys [ ▶ ] or [ ◀ ] to select the number of decimal places, and then [ ▲ ] or [ ▼ ] to change the value.
2 Confirm the entry by pressing [OK] and select Flow source .
3 Using the arrow keys [ ▶ ] or [ ◀ ] , select pump or valve (if you are using an external carrier gas) and confirm by pressing [OK] .
4 Close the Parameters menu by pressing [QUIT] .
5 Press the [FLOW] key.
The gas flow is switched on. The flow rate gradually approaches the level set.
NOTICE
The carrier gas should emerge from the tip of the heating tube into the working medium as tiny bubbles and be spun there forcefully. Alter the stirring rate if necessary. The tip of the tubing should be immersed until it reaches the bottom of the KF titration cell and be rotated against the vessel wall.
Allow the system to condition for a few minutes and swirl the KF titration cell now and again in order to remove the last traces of moisture. Check whether the set oven temperature has been reached.
■■■■■■■■ 31
5.2 Determining the blank value
■■■■■■■■■■■■■■■■■■■■■■
NOTICE
The signal drift on the KF titrator should be less than 20 µg/min (or
µL/min) and remain stable for at least 2 minutes before the blank value determination can be started.
5.2 Determining the blank value
Because the surface of the sample vessels and the air enclosed in the vessel could contain moisture, it is imperative that three to five blank value determinations be carried out. One determination with an empty, tightly sealed sample vessel is carried out in each case. The result is calculated as the quantity of water expelled from the sample vessel. The titration parameters used for a regular sample must also be applied to blank value determinations. Select a relative stop drift of 5 µg/min (or µL/min) as the stop criterion.
Preparing the blank
The system should already be conditioned. A conditioning vessel must be in the oven at this stage and the needle must already be inserted.
1 Seal three to five sample vessels tightly with the crimping tongs.
2 Check that the oven temperature and the gas flow rate correspond to the values which have been set.
3 Wait until the working medium in the KF titration vessel has been conditioned and the signal drift is stable.
Starting the blank value determination
1 Start the determination by pressing the [Start] key on the KF titrator.
2 Use both hands to move the guide head of the 860 KF Thermoprep upward while pressing the fixing lever.
Refer to the illustrations on
ff.
3 Remove the conditioning vessel from the oven with the crimping tongs.
32 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
5 Performing a determination
NOTICE
Caution!
The vessel is hot.
4 Place the blank into the oven using the crimping tongs.
5 Use both hands to move the guide head downward and insert the needle into the blank.
Waiting for the determination
1 Wait until the determination is finished. Conditioning will then be resumed automatically.
2 Repeat the blank value determination with the next blank.
5.3 Determining samples
The blank value must be subtracted from the calculated consumption when determining a sample.
Preparing the KF titrator
1 Load a suitable method.
2 Start the conditioning by pressing the [Start] key.
Preparing the sample
1 Weigh the sample into a sample vessel and seal it tightly with the crimping tongs.
2 Check that the oven temperature and the gas flow rate correspond to the values which have been set.
3 Wait until the working medium in the KF titration vessel has been conditioned and the signal drift is stable.
860 KF Thermoprep
■■■■■■■■ 33
5.3 Determining samples
■■■■■■■■■■■■■■■■■■■■■■
Starting the sample determination
1 Start the determination by pressing the [Start] key on the KF titrator.
2 Use both hands to move the guide head of the 860 KF Thermoprep upward while pressing the fixing lever.
Refer to the illustrations on
ff.
3 Remove the blank from the oven with the crimping tongs.
NOTICE
Caution!
The vessel is hot.
4 Place the sample vessel into the oven using the crimping tongs.
5 Use both hands to move the guide head downward and insert the needle into the sample vessel.
Waiting for the determination
1 Wait until the determination is finished. Conditioning will then be resumed automatically.
2 Repeat the determination with the next blank.
34 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
6 Operation and maintenance
6 Operation and maintenance
6.1 General
The 860 KF Thermoprep requires appropriate care. Excess contamination of the instrument may result in malfunctions and a reduction in the service life of the sturdy mechanical and electronic components.
Severe contamination can also have an influence on the measured results.
Regular cleaning of exposed parts can prevent this to a large extent.
Spilled chemicals and solvents must be removed immediately. In particular, the power plug should be protected from contamination.
6.2 Care
■
■
■
■
■ Replace exhausted molecular sieve promptly. You should replace the molecular sieve as soon as increased drift values appear in the Karl
Fischer cell.
Check all tubing connections regularly for leaks.
Rinse the tubing connections from time to time. Carefully dry the tubing afterward. The tubing must be replaced after prolonged usage.
Replace the dust filter once a year.
Clean the sample holder regularly.
WARNING
Clean the oven only with the instrument switched off and while it is cold.
860 KF Thermoprep
■■■■■■■■ 35
7.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
7 Troubleshooting
7.1 Problems and their solutions
Problem
The drift is very high during conditioning.
The titration time is too long.
Cause
Molecular sieve of drying flasks and/or the titration cell exhausted.
The titration cell is leaking
The sample is non-homogenous
Remedy
Replace molecular sieve.
Check seals. If necessary, replace.
Reduce the size of the sample before weighing in.
The results are spread widely.
Selected gas flow is not achieved.
Not the entirety of the moisture expelled.
Select more stringent switch-off criteria on the
KF titrator: lower stop drift, higher switch-off delay time.
Tubing connections leaking Inspect tubing and replace as necessary.
Molecular sieve of drying flasks exhausted.
Condensate in the heating tube.
Replace molecular sieve.
Gas flow too high.
The sample is non-homogenous
System is leaking
■
■
■
Dry the tubing.
Reduce the gas flow.
Perhaps lower the oven temperature.
Reduce the gas flow.
Reduce the size of the sample before weighing in.
Check all tubing connections and drying flasks for leaks.
Adjust gas pressure (max. 0.3 bar).
Pressure from external gas supply either too low or too high.
Pump defective Contact a service technician.
36 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
8 Appendix
8 Appendix
8.1 Practical notes
Carrier gas selection
If the hot sample is sensitive to air or oxygen (decomposition) and if it releases substances that disrupt the KF reaction, then nitrogen (N
2
) should be used as the carrier gas.
Temperature setting
The temperature selected should be as high as the sample allows (high temperature = shorter analysis time). The sample may not, however, decompose. It must only release water and not any oxidizable substances.
The temperature displayed refers to the temperature in the heating block and not to the sample temperature. Depending on the vial size used, the gas flow and the temperature setting, the actual temperature of the sample can deviate up to 10%.
Gas flow
Once the needle is inserted into the sample vessel until it stops, the gas flow moves through the sample and then through the outlet needle into the titration vessel where the moisture is titrated. If the needle is not permitted to immerse into the sample, then a longer needle holder can be used.
The gas flow should be kept as low as possible. Particularly in the case of very moist samples, care must be taken to ensure that water is not released too quickly and in excessively large quantities into the titration vessel. In addition, there is the danger of condensate formation in the heating tubing. The working medium in the cell must be able to absorb the moisture that is expelled without delay. Normally, a flow rate of 40 to
60 mL/min is optimal.
The tolerances for gas flow measurement depend on the system: They depend on the pressure, temperature, measuring system and the gas itself, and a deviation of 20% is possible.
The gas flow measurement is calibrated with air; other gases deviate more or less strongly from that.
If required, the flow rate of the gas can be checked using a flow measuring instrument at the local pressure and current temperature. An individual measuring factor can be entered with the Meas. factor parameter so that the corrected gas flow is displayed. At the same time, however, other must be selected for the Gas type parameter.
860 KF Thermoprep
■■■■■■■■ 37
8.2 Literature
■■■■■■■■■■■■■■■■■■■■■■
Extraction time
An extraction time of min. 5 minutes should be set on the titrator in order to prevent the titration from being canceled before the sample has released its water.
Conditioning the system
The system must be conditioned with an empty, sealed sample vial (conditioning vial) before a determination is made.
8.2 Literature
■
■
E. Scholz, Hydranal
®
Manual , Riedel-de Haën
P. Bruttel, R. Schlink, Water Determination by Karl Fischer Titration
Monograph, Metrohm, 2006, Order no. 8.026.5013
38 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
9 Technical specifications
9 Technical specifications
9.1 Oven
Temperature range
Accuracy
Correction range
Heating cartridges performance
Heating rate
Cooling rate
50 - 250 °C
±3 °C
–10 - +10 °C
165 W (typical)
Dependent on supply voltage
Typically 15 °C/min (at 80 - 180 °C, 230 V)
Dependent on temperature, supply voltage, sample amount and vessel dimensions
Typically 9 °C/min (at 180 - 80 °C)
Dependent on temperature, sample amount and vessel dimensions
9.2 Gas flow
Flow range 10 - 150 mL/min
Under normal conditions, calibrated with air at approx. +25 °C and
1013 mbar.
Deviations of gas amount per time unit of ±20% are possible.
9.3 Outlet heater
Socket connection
Typical tubing temperature
U = 16 ±1 V
I ≤ 0.8 A approx. 50 °C
860 KF Thermoprep
■■■■■■■■ 39
9.4 Ambient conditions
■■■■■■■■■■■■■■■■■■■■■■
9.4 Ambient conditions
+5 °C - +45 °C, at max. 80% humidity, non-condensing Nominal function range
Storage
Altitude / Pressure range
Pollution degree
+5 - +45 °C
Max. 2,000 m.a.s.l. / min. 700 mbar
2
9.5 Energy supply
100 - 120 / 220 - 240 V (±10%) Nominal voltage range
Nominal frequency
Power consumption
Protection
Overvoltage category
50 / 60 Hz
200 W
2 ATH
II
9.6 Reference conditions
25 °C (±3 °C) Ambient temperature
Relative humidity
Supply voltage
≤ 60%
230 V
40 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
9.7 Dimensions
Width
Height
0.44 m
0.18 m (without support rod)
0.50 m (with support rod)
0.23 m
8.17 kg
Depth
Weight (without accessories)
Material
Housing
Oven covering
Operating unit
Cover and bottom: Steel sheet, stove-enameled
PTFE
Polycarbonate / Acrylonitrile butadiene styrene (PC/ABS)
9 Technical specifications
860 KF Thermoprep
■■■■■■■■ 41
■■■■■■■■■■■■■■■■■■■■■■
10 Accessories
Up-to-date information on the scope of delivery and optional accessories for your product can be found on the Internet. You can download this information using the article number as follows:
Downloading the accessories list
1 Enter https://www.metrohm.com/ into your Internet browser.
2 Enter the article number (e.g. 860 ) into the search field.
The search result is displayed.
3 Click on the product.
Detailed information regarding the product is shown on various tabs.
4 On the Included parts tab, click on Download the PDF .
The PDF file with the accessories data is created.
NOTICE
Once you have received your new product, we recommend downloading the accessories list from the Internet, printing it out and keeping it together with the manual for reference purposes.
42 ■■■■■■■■
860 KF Thermoprep
■■■■■■■■■■■■■■■■■■■■■■
Index
Index
6.2627.000 oven insert ............ 13
A
Arrow button ........................... 22
Arrow key .......................... 24, 25
Assembling
Drying flasks ....................... 17
B
Basic settings ............................ 27
Blank value ............................... 32
Brightness ................................ 23
C
Cable holder ......................... 7, 15
Carrier gas ................................ 37
Colon ....................................... 24
Conditioning ............................ 28
Conditioning vessel .................. 29
Connect
Power grid ............................ 8
Connector
Air/nitrogen .................... 7, 19
Outlet heating ...................... 7
Crimping tongs. ........................ 29
D
Display ..................................... 23
Drying flask .................... 6, 17, 18
Dust filter ............................. 7, 16
E
Electrostatic charge .................... 4
Error message ........................... 24
External gas connector ............. 19
Extraction time ......................... 38
F
Fan ............................................. 7
Fixing lever ........................... 6, 30
Flow rate .................................. 23
Flow rate error ......................... 17
Flow source .............................. 26
G
Gas flow ....................... 23, 26, 37
Gas inlet ..................................... 6
Gas outlet .................................. 6
Gas supply ................................ 26
Gas type ................................... 26
Guidance rod ............................. 6
Guide head .......................... 6, 30
H
Heating tube ............................ 19
Heating tubing ......................... 15
I
Icon ......................................... 23
Injection needle ........................ 14
Inlet
Air pump .............................. 7
Inlet filter .................................. 16
K
Keys ......................................... 22
KF titration cell ................... 19, 28
Coulometric ....................... 19
Volumetric .......................... 20
L
Lock keys .................................. 27
M
Main dialog .............................. 22
Maintenance ............................ 35
Molecular sieve ........................ 17
Mount
Dust filter ........................... 16
Heating tubing ................... 15
Support rod .......................... 9
Tubings .............................. 18
Mounting
KF titration cell ................... 19
Needle ............................... 13
Sample holder .................... 12
N
Needle ..................................... 13
Needle holder ........................... 14
Numerical input ........................ 25
O
Outlet heater ............................ 15
outlet needle ............................ 14
Oven module ............................. 6
Oven temperature .............. 23, 25
P
Parameters ............................... 25
Personnel safety ......................... 4
Power connection .................. 8, 9
S
Safety instructions ...................... 3
Sample determination .............. 33
Sample holder ...................... 6, 12
Sample vessel ............................. 6
Selection .................................. 25
Selection list ............................. 25
Serial number ............................. 7
Service ....................................... 3
Setup ....................................... 27
Signal drift ................................ 32
Solvents ..................................... 5
Status ....................................... 23
Supply voltage ............................ 4
Support rod ................................ 9
T
Temperature ................. 23, 25, 37
Temperature correction ............ 27
Tubings .................................... 18
Type plate ................................... 7
V
Valve ........................................ 26
860 KF Thermoprep
■■■■■■■■ 43
advertisement