WATT/Leakage METER

Model : DW-6160

Your purchase of this

WATT/Leakage METER marks a step forward for you into the field of precision measurement.

A l t h o u g h t h i s

WATT/Leakage METER is a complex and delicate instrument, its durable structure will allow many years of use if proper operating t e c h n i q u e s a r e developed. Please read t h e f o l l o w i n g instructions carefully and always keep this manual within easy reach.

OPERATION MANUAL

Caution Symbol

Caution :

* Risk of electric shock !

Caution :

* Do not apply the overload voltage, current to the input terminal !

* Remove test leads before open the battery cover !

* Cleaning - Only use the dry cloth to clean the plastic case !

Environment Conditions

* Installation Categories II 600V.

* Pollution Degree 2.

* Altitude up to 2000 meters.

* Indoor use.

* Relative humidity 80% max.

Equipment protected throughout by double insulation or reinforced insulation.

TABLE OF CONTENTS

1. FEATURES................................................................. 1

2. SPECIFICATIONS....................................................

2

2-1 General Specifications............................................ 2

2-2 Electrical specifications...........................................3

3. FRONT PANEL DESCRIPTION..................................... 6

3-1 Display..................................................................6

3-2 Backlight button................................................. 6

3-3 Record button ( MAX. / MIN. )............................... 6

3-4 Data hold button.................................................. 6

3-5 Power & function rotary switch ..........................

6

3-6 ACV / DCV & Ω input terminal.............................. 6

3-7 Load terminal....................................................... 6

3-8 Power source terminal......................................... 6

3-9 Battery Cover / Compartment.................................6

3-10 Stand................................................................ 6

3-11 RS-232 output terminal........................................ 6

3-12 DC 9V power adapter input socket........................ 6

4. MEASURING PROCEDURES.......................................... 7

5. MAINTENANCE........................................................... 8

6. RS232 PC SERIAL INTERFACE.................................... 9

7. BATTERY REPLACEMENT........................................... 10

8. THE ADDRESS OF AFTER SERVICE CENTER..................11

1. FEATURES

* Professional WATT meter with Leakage current tester digital display , battery operated .

* LSI - circuit provides high reliability and durability .

* Measurement

WATT ( AC ) : 2500W x 0.1 W / 1 W.

Leakage current ( AC mA ) : 80.00 mA x 0.01 mA.

ACV : 600.0 V x 0.01 V / 0.1 V.

ACA : 10.00 A x 0.001 A / 0.01 A.

DCV : 600.0 V x 0.01 V / 0.1 V.

Ω : 20 k Ω x 0.01 k Ω / 0.001 k Ω / 1 Ω .

* Low Watt measurement , 999.9 Watt x 0.1 W.

* True Power and Wide range 0 to 2500 Watt.

* True RMS measurement of ACV / ACA.

* Leakage detect ( AC mA ) detection.

* Continuity beeper.

* Super large LCD , Dual display at same time.

* Full function Auto range.

* Memory Record ( Max. , Min. ) , Data Hold.

* RS232 / USB Computer interface.

* Built-in low battery indicator.

* Built-in backlight easy to readout .

* Power : DC 1.5V (UM-3 , AA ) x 8 PCs or DC 9V adapter in.

1

2. SPECIFICATIONS

2-1 General Specifications

Circuit

Display

Custom one-chip of microprocessor

LSI circuit.

Large LCD display Max. reading 9999.

LCD size :90.5 x 64.5 mm.

Measurement

Dual value display with backlight.

ACV

DCV

WATT

0 to 600.0 V

0 to 600.0 V

0 Ω to 20.00 k Ω

0.1 W to 2500 W

Leakage 0.01 m A to 80.00 m A current

ACA 0.01 A to 10.00 A

" - - - - " mark indication .

Over input

Polarity Automatic switching , "-" indicates reverse polarity.

Zero Adjustment Automatic adjustment.

Sampling Time Approx. 1.5 second.

Data Hold

Data output

Freeze the display reading.

Memory Recall Maximum & Minimum value.

RS232 / USB PC Computer interface .

* Connect the optional RS232 cable

UPCB - 02 will get the RS232 plug .

* Connect the optional USB cable

USB - 01 will get the USB plug .

℃ ℃ ℉ ℉

Operation

Temperature

Operation

Humidity

Less than 80% RH.

2

Power Supply

Power

Consumption

Weight

Dimension

Standard

Accessories

Included

Optional

Accessories

Alkaline or heavy duty DC 1.5 V battery

( UM3, AA ) x 8 PCs, or equivalent.

* A

adapter is optional ).

Approx. DC 33 mA

883 g / 1.95 LB

224 x 125 x 65 mm

( 8.8x 4.8 x 2.5 inch )

Instruction manual.................1 PC

Test lead ( Red & Black )........1 pair

* AC/DC 9V adapter, AP-9VA.

* USB cable , USB - 01.

* RS232 cable , UPCB - 02.

* Data Acquisition software,

SW-U801-WIN.

* Hard carrying case , CA-08.

AC / DC VOLTAGE

Range Range 1 0.01V to 99.99V

Range 2 100 V to 600V

Resolution

Accuracy

Range 1 0.01 V

Range 2 0.1 V

ACV

DCV

± (1 %+5d)

± (0.8 %+5d)

* Measuring Signals come from the front panel terminals.

* Auto range.

* Max. input voltage : AC 600 V, DC 600 V.

* ACV accuracy is test under input signal is sine wave,

50/60 Hz

* ACV frequency response is from 40 to 400 Hz.

* ACV is true rms measurement..

3

OHMS ( Resistance )

Range Resolution Accuracy

0 to 999 Ω 1 Ω

1.000 to 9.999 K Ω 0.001 K Ω

10 to 20 K Ω 0.01 K Ω

± (1 % + 1d )

± (1 % + 1d )

± (1 % + 1d )

* Auto range.

* Continuity beeper : < 4 Ω .

* Overload Rating : AC / DC 600V at 20 second Max

WATT ( true power )

Range Resolution Accuracy

1000 W 0.1 W

1001 to 2500 W 1 W

± (1.5 % + 5d )

± (1.5 % + 5d )

* 0.1 W resolution : Input voltage < 120 AVC or

Input current < 1 ACA.

Beyond above input, the resolution will still be 1 W.

* Measuring Signals come from the top power plug input..

* Accuracy is test under input signal is sine wave, 50/60 Hz

* ACV, ACA frequency response is from 40 to 400 Hz.

* Max. input voltage : AC voltage 250 V, AC current : 10 A.

Leakage ( AC mA )

Range

0 to 80 mA

Resolution

0.01 mA

Accuracy

± (1 % + 5d )

* The leakage current that sense between the " Hot line " and the "

Earth " of the measuring installation that connect to the output "

Power plug "

4

V, A ( true rms )

Range

ACV 250 V

ACA 10 A

Resolution

0.01 V/0.1 V

0.001 A/0.01A

Accuracy

± (1 % + 5d )

± (1 % + 3d )

* ACV, 0.01 V resolution is valid from 0.01 V to 99.99 V.

ACV, 0.1 V resolution is valid from > 100.0 V.

* ACA, 0.001 A resolution is valid from 0.001 A to 1.200 A.

ACA, 0.01 A resolution is valid from > 1.2 A.

* Measuring signals come from the top power plug input

( power source ).

* Auto range.

* ACV, ACA accuracy is test under input signal is sine wave,

50/60 Hz

* ACV, ACA frequency response is from 40 to 400 Hz.

* ACV, ACA is true rms measurement..

* Max. input voltage : AC voltage 250 V, AC current : 10 A.

5

3. FRONT PANEL DESCRIPTION

Fig. 1

3-1 Display

3-2 Backlight button

3-3 Record button ( MAX. / MIN. )

3-4 Data hold button

3-5 Power & function rotary switch

3-6 ACV / DCV & Ω input terminal

3-7 Load terminal

3-8 Power source terminal

3-9 Battery Cover / Compartment

3-10 Stand

3-11 RS-232 output terminal

3-12 DC 9V power adapter input socket

6

4. MEASURING PROCEDURE

4-1 ACV measurement procedure

1)Rotate the function switch to the " ACV " position and

Power ON the meter.

2)Connect Red test lead to the " V Ω " input terminal and

Black test lead to " COM " input terminal.

3)Use the test lead measurement unknown AC voltage.

4)Read the Display. The show value will correspond to the position selected. If the Display shows " ---- ", it indicates an out-of-range measurement.

4-2 DCV measurement procedure

1)Rotate the function switch to the " DCV " position and

Power ON the meter.

2)Connect Red test lead to the " V Ω " input terminal and

Black test lead to " COM " input terminal .

3)Use the test lead measurement unknown DC voltage.

4)Read the display . The show value will correspond to the position selected. If the Display shows " ---- " it indicates an out-of-range measurement.

Remark :

* Observe polarity when measuring DC voltage.

4-3 Resistance measurement procedure

1)Rotate the function switch to the Ω position and Power

ON the meter.

2)Connect red test lead to the " V Ω " input terminal and black test lead to " COM " input terminal .

Connect test lead to the resistance.

7

3)Read the display . The show value will indicate correspond to the position selected. If the display shows " ---- " it indicates an out-of-range measurement.

4)When resistance < 4 Ω the buzzer will continuity beeper .

4-4 Watt measurement procedure

1)Rotate the function switch to the " WATT/Leakage " position.

2)Make sure your equipment work voltage, such as 110 V or 220 V.

3)Plug in the power source to " AC POWER INPUT " terminal, it locate the left position of Top Panel.

4)Plug in the equipment power line to the " LOAD " Terminal, it locate the right position of Top Panel.

5)Turn on the equipment power switch.

6)It can read the Power Consumption value on the top LCD

Display. At the same time. If your equipment exist current leakage. It will show the leakage current value on the bottom LCD Display.

4-5 V ( voltage ) A ( current ) measurement procedures

When your equipment is on working. You can rotate the function switch to V A Position. The upper value will show working voltage. The bottom value will show the current consumption value.

5. MAINTENANCE

1)The meter is intended for measuring the TRUE RMS

POWER VALUE, at the same time it also can measure the leakage current value to verify if the measurement equipment is under safety condition or not.

2)When measuring POWER VALUE, the working voltage of measurement equipment must be checked before turn on the equipment.

8

3)Instruments used in dusty environments should be stripped and cleaned periodically.

4)Do not leave the instrument exposed to direct heat from the sun for long periods.

5)Before removing the battery compartment cover, ensure that the instrument is disconnect from any circuit and the power switch is on the OFF position.

6. RS232 PC SERIAL INTERFACE

The instrument has RS232 PC serial interface via a 3.5

mm terminal ( 3-11, Fig. 1 ).

The data output is a 16 digit stream which can be utilized for user's specific application.

A RS232 lead with the following connection will be required to link the instrument with the PC serial port.

Meter PC

(9W 'D" Connector)

Center Pin..........................Pin 4

(3.5 mm jack plug)

Ground/shield.......................Pin 2

2.2 K resistor

Pin 5

The 16 digits data stream will be displayed in the following format :

D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0

9

Each digit indicates the following status :

D0 End Word

D1 & D8 Display reading, D1 = LSD, D8 = MSD

For example :

If the display reading is 1234, then D8 to

D1 is : 00001234

Decimal Point(DP), position from right to the D9

D10 left

0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP

Polarity

0 = Positive 1 = Negative

D11 & D12 Annunciator for Display

D13

D14

D15

ACV = 50 DCV = 34 K Ω = 39

Ω = 38 W = 47 m A = 53

ACA = 52

When send the upper display data = 1

When send the lower display data = 2

4

Start Word

RS232 FORMAT : 9600, N, 8, 1

Baud rate 9600

Parity

Data bit no. 8 Data bits

Stop bit

No parity

1 Stop bit

7. BATTERY REPLACEMENT

1)When the left corner of LCD display show " " , it is necessary to replace the battery, However in -spec.

measurement may still be made for several hours after

Low Battery Indicator appears before the instrument become inaccurate.

10

2)Open the screw of "Battery Cover "by screwdriver, then move the batteries.

3)Replace with 1.5V x 8 PCs batteries and reinstate the cover.

8. THE ADDRESS OF AFTER SERVICE

CENTER

3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]

MRC.

7

.1

8

11

advertisement

Related manuals

Frequently Answers and Questions

What is the maximum input voltage for AC and DC measurements?
The maximum input voltage for both AC and DC measurements is 600V.
Can the MRC DW-6160 measure AC mA?
Yes, it can measure AC mA with a range of 0 to 80 mA and a resolution of 0.01 mA.
What is the accuracy of the ACV measurement?
The accuracy of the ACV measurement is ± (1% + 5d).
What is the resolution of the resistance measurement?
The resolution of the resistance measurement varies depending on the range: 1Ω for the 0 to 999 Ω range, 0.001 kΩ for the 1.000 to 9.999 kΩ range, and 0.01 kΩ for the 10 to 20 kΩ range.
Can the MRC DW-6160 measure true power?
Yes, it can measure true power with a range of 1000W to 2500W and a resolution of 0.1W.