advertisement
▼
Scroll to page 2
of
34
INSTRUCTION AND SERVICE MANUAL TBE/A! OXYGEN ANALYZERS A MODELS 350, 351, 352, 353, 354 & 355 CUSTOMER ORDER NOG 0 ~~ SALES ORDER No. Lo —[ reledyne Analytical Instruments 16830 Chestnut Street City of Industry, CA 91749 — 1580 Phone: (626) 934— 1500, Fax: (626) 961 — 2538 TWX (910)584 — 1887 TDYANLY COID Web: www.teledyne —ai.com 5 TABLE OF CONTENTS INTRODUCTION. . .ecocccnorocrocerecoreoce acen. rs a 0000020000 DESCRIPTIONS 2.1 Models and Sensors.....eíereeoorcscsrreorecacreseroaaoroo 2.1.1 Model 350 * # Y & + XA AAA + + 0 + 4 # * 2.1.2 Model 351......e.recroo. ese esse e encroocrerreoo 2.1.3 Model 352... . г... ... y oscororesorovaroerosec e 2.1.4 Model 353......e...eo. e orceocrocsrerovrooreraeco 2.1.5 Model 354.......... + rscorcovorecanranaerercerco 2.1.6 DENSOFS....cocomoarascaracorcreoaracacooorado 2.2 Controls and Display. ........ 0 060400.000000080 08080008 . 2.2.1 Calibrate......we.esoocsoosrererccero eve us 2000 2.2.2 Off/On a « # 4 + + 5 + + + + ¥ 8% WF BREF EREDAR PED FEEDS EES 2.2.3 Alarm Setpoints (Models 352 and 354)........ 2.2.4 Alarm Indication...evveveevrencnneeenns .... 2.2.5 Meter Display....... 0050000800020 008 ceva .. 2.3 Sensor Attributes and Circuit Requirements.......... .. 2.3.1 Sensor AttributeS...0..0.000000600000400000000080 « 2.3.2 Circuit Requirements... cvveerecevrnevsoonses 2.4 Individual Circuits and Components....._.ec-o.roesecocooo 2.4.1 Sensor Amplification Circuit.......eomceccooo 2.4.2 Reference Voltage Circuit. ...o..ecoe.ecocvcaso 2.4.3 Alarm Circuit - Standard Configurations..... 2.4.4 Alarm Logic Circuit. ...ce.ec.eocorcoresoaraceneo 2.4.5 The Meter, ......ooce.cocenocooenoccearo escoreco $ 2.4.5.1 Analog Meter.......eooesocrcoroeo. . 2.4.5.2 Digital Meter......e.eoe.e. cose cans 2.4.6 Power Supply. ...o.ocorcasoocarerarermarcarmarao INSTALLATION — BATTERY AND SENSOR. ...eocresoro. ere 500000 3.1 Battery Requirements. ....eoomeoscosonococsoocoacoaooacono 3.2 Battery Installation (Models 351, 352, 353, 354).. 3.3 Sensor Installation for B-6 Cell........ es see one 00000 3.4 Sensor Installation for B-5 Cell. .....r.eerecsooreocaooo 4 OPERATING INSTRUCTIONS. ......uueurssnnnennnsnnnsssnnenn,, 10 Model 350s se tte cre ana 10 Model ID 10 Model Usa TS 10 Model 353. e eerencaroosacan cena oarerenreCeaanone o 11 4.5 Model BHO ecco 12 + » > Ja 0 > Y DN pe 5 SERV ТСК INSTRUC TIONS т тов нет теле * + + # # + + + 5 5 5 6 à + + + EEE. 13 5 * 1 Troubleshoot ing. =" EEES A Eo. "70. 0409 4 04 4 0 9 48 чево тени 13 5. 1 * 1 Preliminary Checks > E 0 4 4 dada 0 04 0 4 4 4 4 X 4 8 4 13 5.1.2 Measurement Circuit Checks......e..e.eeec0... 13 5.1.3 Alarm Circuit Checks (Models 351, 352, 352 and 334). ......cccrorccoooococorarecocacove, 13 - 5.1.4 Meter Circuit Checks..........ee.o.conancoo. 14 5.2 Repair/Replacement...................e..0ecccom ree, 14 5.2.1 - Battery Replacement...........eeeeconcooce., 14 5.2.2 Sensor Replacement............eeoeococooooo. 15 5.2.3 Э1ваввешЬ1у. 55. ооооьовооьо оо очьовавово, 15 5.3 Manufacturing Test Procedures, ..........eooaccooocooa, 15 6 MODEL 355 INSTRUCTIONS. ............e.eecocneroconocenaoea 21 MATERIAL SAFETY DATA SHEET | - SPARE PARTS LIST REFERENCE DRAWINGS INTRODUCTION This manual provides operating instructions and service for the 350 Series of oxygen analyzers. information Models 350, 351, 352, 353, 354, and 355 provide continuous and speci- fic monitoring of oxygen. Each model is a fully independent unit designed to monitor the oxygen content of an atmosphere to which its sensor is exposed, Model 355 is quite different in physical арреаг- ance from the other instruments in the series, For this reason, separate coverage for Model 355 appears in Section 6. | Models 350, 351, 352, 353, and 354 employ one of Teledyne Analytical Instruments” patented electrochemical sensors (Cell Class B-6 or B-5), which is acted upon by oxygen in the sample, to produce an electric current, the strength of which is determined by the amount (concentra- tion) of oxygen present. Typical features of standard instruments are: Model Number 350 351 352 353 354 Feature: Power -—- None 4-0 4-0 29V 2-3 Batteries Sensor MICRO—FUEL CELL Alarms None Fixed LO, HI, Fixed LO, HI, LO, Visual, Visual, LO, Visual, Visual, Audio Audio, Audio, 30 Audio, Sec, Audio 30 Sec. Silence Audio Silence Display Analog Analog Analog Digital Digital Meter Meter Meter LCD LCD The sensors sample gaseous mixtures for oxygen by diffusion or, if fitted with special adaptors, gases confined in closed systems or containers. Qualified personnel who service the TAI analyzers must be able to read and follow schematics and know the basic functions and character- istics of operational amplifiers. The only equipment required for troubleshooting the analyzer is a Digital Volt Meter (DVM); however, a way to simulate the sensors may be useful. Section 5.1 contains suggestions for troubleshooting. 2 DESCRIPTIONS 2.1 Models and Sensors 2.1.1 Model 350 The Model 350 oxygen analyzer is a simple, lightweight unit equipped with an easy-to-read analog meter with 1% oxygen gradua- tions. Its rugged case will withstand industrial use. The cable of the sensor holder plugs conveniently into a receptacle in the case. The oxygen sensor directly drives the meter of the Model 350 without batteries or electronic circuitry. 2.1.2 Model 351 Model 351 provides the features of the Model 350 with the addi- tion of a fixed low-level alarm to indicate when the oxygen level is below the setpoint. Model 351 provides both audible and visual (indicator light) alarms. | 2.1.3 Model 352 Model 352 provides the features of the Model 351 with the addi- tion of fully adjustable high- and low-level alarms to indicate when the oxygen level is either above or below the selected setpoints. The setpoints for the alarm circuits are adjusted by digital pushbutton potentiometers that indicate the setpoint directly in numerals. Model 352 provides both audible and visual (indicator light) alarms. 2.1.4 Model 353 Model 353 contains a digital Liquid Crystal Display (LCD) instead of the analog (pointer-type) meter. Because the digital display is smaller, the Model 353 case is more compact. The digital display produces alarm and battery level condition messages on the display and provides a direct reading of the oxygen level in numerals. Model 353 provides a fixed low-level alarm that is both audible and visual (indicator light). Also, the Model 353 provides a pushbutton to silence the audible alarm for 30 sec- onds, after which it reactivates automatically, unless the alarm condition has been corrected. 2.1.5 Model 354 Model 354 provides the features of Model 353 with the addition of fully adjustable high- and low-level alarms to indicate when the oxygen level is either above or below the selected setpoints. The setpoints for the alarm circuits are adjusted by digital pushbutton potentiometers that indicate the setpoint directly in numerals. | 2.2 2.1.6 Sensors The standard sensor for these instruments is Teledyne’s patented Micro-Fuel Cell, The B-6 cell, used in Model 350 only, resides in a durable, corrosion-resistant holder. The holder connects to the analyzer with a coiled cable plugged into a receptacle. The cell connects to the analyzer electronics via spring-loaded con- tacts on one end, pressed against mating contacts in the cell holder. A screw-on cap holds the cell in its proper position within the holder. The B-5 electrochemical sensor assembly is used in all other models of the 350 series (Models 351, 352, 353, and 354). The cell and probe combine to form a single replaceable unit. A coiled cord plugs into both the sensor assembly and the analyzer. Replacement of the cell and probe is’ simple and requires no tools. The Micro-Fuel Cell is a self-contained electrochemical trans— ducer that requires no maintenance. It provides stable and reliable readings throughout its useful life, after which it is easily replaced. Because the Micro-Fuel Cell generates no signal in the absence of oxygen, no zero standardization is required. When practical, the analyzer should be calibrated to the oxygen content of air; this procedure is described in detail later. Controls and Display 2.2.1 Calibration The face of each instrument contains a control potentiometer to adjust for variations in sensor output. When the sensor is exposed to a known oxygen concentration, the "Calibrate!" knob is turned to adjust the meter indication to the known concentration. The known oxygen concentration of air (20.9%) is used as the standard for this adjustment whenever it is practical. 2.2.2 OFf/On Use the Off/On switch to turn Models 351, 352, 353, and 354 on and off. Model 350 does not use batteries or powered electronic circuits, so it does not require an Off/On switch. Models 351 and 352 incorporate a "Batt, Test" position into the Off/On switch for checking the condition of their batteries against a scale indication on their meters. Battery level condition for Models 353 and 354 appears on the LCD display and will be des- cribed later. 2.2.3 Alarm Setpoints (Models 352 and 354) Models 352 and 354 are equipped with one HI and one LOW oxygen level alarm. The alarms for Models 352 and 354 are fully adjustable by means of digital pushbutton potentiometers. By 2.3 pressing the minus (~) pushbutton below one of the numerals on the face of the control, that numeral is decreased by one count. Conversely, pressing the plus (+) pushbutton above a numeral adds one count to that numeral. The setpoint indicators represent percent oxygen settings in increments of 1% (right-hand numeral) and 10% (left-hand numeral). 2.2.4 Alarm Indication Models 351 and 353 have only low-level alarms. Models 352 and 354 have fully adjustable high- and low-level alarms. Models 351, 352, 353, and 354 provide both audible and visual alarm indication in the event that the measured oxygen level deviates from the alarm setpoint. If desired, on Models 352 and 354 you can silence the audible alarm for periods of 30 seconds each by pressing the "Audio Silence" pushbutton on the front panel. The audible alarm will sound again after the 30-second time period, unless the alarm condition has been corrected. The visual alarm indicator continues to warn of the alarm condition until it is corrected, regardless of the state of the audible device. 2.2.5 Meter Display Models 350, 351, and 352 come with a very accurate analog (pointer-and-scale type) meter. Standard instruments contain a mirrored scale on the face of the meter, so the operator can align the pointer accurately with its reflection on the mirrored scale and avoid reading errors due to parallax. The meters on Models 351 and 352 include a Battery Check scale that indicates the battery condition while the power switch is in the "BAT TEST" position. Models 353 and 354 provide oxygen level indication in numerals on the digital LCD rather than the pointer-scale type analog meter. Also, the digital readout provides alarm and low battery messages in words when the indicated condition exists, Sensor Attributes and Circuit Requirements 2.3.1 Sensor Attributes Each of the instruments uses the Micro-Fuel (galvanic-type) cell. A thermistor/resistor circuit compensates for temperature-related effects of the sensor. The B-6 Micro-Fuel Cell uses a probe assembly with built-in thermistor circuitry. The B-5 Micro-Fuel Cell Assembly contains a sensor and the thermistor circuitry in a single unit. The Micro Fuel Cell does not require an external excitation voltage source; it will independently produce a current propor- tional to the amount of oxygen present in the gas mixture sur- rounding it. 2.4 A typical Micro-Fuel Cell, the B-6, produces 300 microampere (+20%) of current in 100% oxygen. The B-5 sensor assembly produces 100 millivolts (+20%) in 100% oxygen. 2.3.2 Circuit Requirements The current from the Micro-Fuel Cell passes through a network consisting of a thermistor and two resistors within the probe asgembly, and then through a span potentiometer (span pot) in the analyzer unit. This converts the non-compensated cell current into a temperature-compensated voltage. The span pot is used to calibrate the sensor, and an amplifier is used to bring the voltage up to a suitable level for driving the meter and alarm circuitry. Because the B-5 sensor assembly is one unit including the thermistor circuitry, the B-5 produces voltage rather than current output. As an example, refer to the Model 351 Schematic provided in the drawings section. The Micro-Fuel Cell is shown with its compen- sating circuitry connected to the span pot. The slider of the span pot is connected to the non-inverting input of operational amplifier Alc, which amplifies the signal. Offset adjustment for the amplifier is provided by P-1. Individual Circuits and Components Each of the analyzers in this product series contains some of the following features: Sensor amplification circuit Reference voltage circuit Alarm circuit Alarm bypass logic Meter (analog or digital) Power supply Oa 69 ВО This chart lists each feature and indicates the analyzers in which that feature is found. Use the chart for reference when you perform the troubleshooting tests outlined in section 5. section: Used in: 350 351/352 353/354 Sensor amplification circuit NO YES YES Reference voltage circuit NO YES YES Alarm Circuit NO YES YES Alarm Bypass Logic NO NO YES Meter (analog) YES YES NO Meter (digital) NO NO YES Power Supply NO YES YES 2.4.1 Sensor Amplification Circuit See Section II —- "DESCRIPTIONS ~ Sensor Attributes and Circuit Requirements." 2.4.2 Reference Voltage Circuit The circuits for the alarms require a reference voltage to work properly. Reference voltage is provided by a band-gap reference, IM385, which acts like a very precise and stable zener diode, producing 1.25 volts (or 2.50 volts). Some circuits use an amplifier to lower the impedance of the reference and produce a l volt reference for other parts of the circuit. An incorrect reference voltage can be caused by a bad reference IC, by too high a resistance feeding it, or by too low a resistive load on it. Problems with the alarm setpoints may indicate a faulty reference voltage circuit or an improper reference voltage level. 2.4.3 Alarm Circuit - Standard Configurations The available alarm configurations are: Type of alarm Used in 350 351 352 353 354 High - adjustable NO NO YES NO YES Low =~ adjustable NO NO YES NO YES Low - fixed at 19,5% NO YES NO YES NO 30-Second Silence NO NO NO YES YES Battery Condition NO YES YES YES YES In each case an operational amplifier, used as a comparator, determines whether the measured parameter (oxygen level or bat- tery voltage) is greater or less than a preset voltage, The outputs of the comparators are fed to a prescribed logic circuit, then to the alarm output devices (L.E.D., beeper, or LCD) as determined by the alarm configuration. As an example, refer to the Model 351 Schematic in the drawings section. R7 and ZD1 produce the 1.25 volt reference. A fixed portion (0.78 volts) of the 1.25 volts is fed to Aid, which compares that voltage with the oxygen level from Alc; if that level drops below 0.78 volts, the output of Ald goes positive, turns on transistor Ql via D3 and R17, and turns on the LED and the beeper. 2.4.4 Alarm Logic Circuit Models 353 and 354 have logic that allows the beeper to be silen- ced for 30 seconds by pressing the switch on the front of the unit, Comparators A3a, A3b, and A3c can all generate alarms (see sche- matics for Models 351 and 352). A3a is the high alarm, A3b is the low alarm, and A3c is the fixed low alarm. Diodes D5; D6, D7, D10, Dil, and D14 are used to "or" the outputs to the approp- riate display points on the LCD and also into the circuitry that uses Ul. Integrated circuit Ul performs the following functions via the IC pin connections by which they are accessed. Pins 9 and 8 form an oscillator that is disabled by pins 1 and 2. When enabled, it causes the LED to blink via pins 3 and 4. Also, it fires the beeper via D13, but only upon a low output at Ul pins 11 and 10, The low condition at pins 11 and 10 exists under alarm conditions except when the pushbutton is pressed (Models 353 and 354 only), at which time C8 is discharged (and the alarm silenced) for 30 seconds, until CB has charged up again. 2.4.5 The Meter 2.4.5.1 Analog Meter. The analog (pointer-and- scale-type) meters are simply driven via a trimpot from the 0-1 volt output of an operational amplifier. The meter is calibrated by first mechanically zeroing it, next causing the voltage on it to be 1 volt, and finally adjusting the series trimpot to achieve a fullscale reading. 2.4.5.2 Digital Meter. Digital instruments use an integrated analog-to-digital converter that needs no adjustment. The auxiliary messages (Hi Alarm, Lo Alarm, Wait, and Lo Bat) on the ICD are driven by "exclusive or" gates in a separate IC and are enabled by a "high" to the appropriate line, 2.4.6 Power Supply Power is supplied by batteries (Model 350 does not have a power supply). Models 351 and 352 use 4 "C" cells and provide a balan- ced plus and minus 3 volts (nominally) with the center connec- tion of the batteries being the circuit reference (common or ground). Models 353 and 354 use two 9-volt transistor batteries and provide the common, or ground, connection by the use of an amplifier. М Common (ground) connection The output voltage of the amplifier is the same as the voltage at the junction of Ri and RZ. The amplifier can supply as much current as needed for the rest of the circuitry. 3 INSTALLATION - BATTERY AND SENSOR 3.1 3.2 3.3 CAUTION: 3.4 Battery Requirements Model 350: None Models 351 and 352: Four size "C" batteries Models 353 and 354: Two standard 9-volt batteries Battery Installation (Models 351, 352, 353, 354) i. 2. 3. 4. Turn the power switch to the OFF position. Move the sliding back panel to the right to expose the battery compartment. Install the correct batteries, observing proper polarity. Close the sliding back panel. sensor Installation for B-6 Cell Do not scratch, puncture, or otherwise damage the Micro-Fuel Cell membrane. Damage to the membrane will require sensor replacement. NEVER PRESS ON THE SENSING SURFACE; you might damage the sensor. 1. Hemove the new B-6 Micro-Fuel Cell from its protective bag. Remove the shorting clip and tape from the contact plate. Unscrew the holder cap from the sensor holder. Place the Micro-Fuel Cell inside the sensor holder with the cell’s membrane surface facing outward and the contact plate facing inward. Screw the cap back onto the sensor holder. Check to see that the sensor cable is plugged into the receptacle on the front of the analyzer. Sensor Installation for B-5 Cell 1. Remove the new B-5 sensor assembly from its protective bag. Remove the coiled cord from its protective bag. Plug the cord into the sensor assembly. Then, plug the other end of the cord into the analyzer. 4 OPERATING INSTRUCTIONS 4.1 Model 350 4.2 4.3 1. Be sure that the Micro-Fuel Cell sensor is installed properly in the sensor holder. Zu Connect the plug from the sensor holder cable into its receptacle on the front panel of the analzyer. 3, Expose the sensor holder to air (20.9% oxygen) and adjust the "Calibrate" control until the meter needle points exactly to the calibration mark on the meter scale. 4, Model 350 is ready for use. Model 351 1. © Hold the ON/OFF switch down in the BAT TEST position. д, Check the meter for battery condition; replace batteries if necessary. de Be sure that the B-§ electrochemical oxygen sensor assembly has the sensor cable plugged into it. 4, Connect the other end of the plug from the sensor cable into 1ts receptacle on the front of the analyzer and turn the instrument on. 6. Expose the sensor to air (20.9% oxygen) and adjust the "Calibrate" control until the meter needle points exact- ly to the calibration mark on the meter scale. 5, Check the fixed low alarm by adjusting the "calibrate" control until the meter reads below 19.5%. The audible and visual alarms should both activate, 7. Model 351 is ready for use. Model 352 1. Hold the ON/OFF switch down in the BAT TEST position, 2. Check the meter for battery condition; replace batteries, if necessary, | de Be sure that the B-D electrochemical oxygen sensor has the sensor cable plugged into it. 4. Connect the other end of the plug from the sensor cable into its receptacie on the front of the analyzer and turn the instrument on. 10 4.4 LA 8. Calibrate the instrument to a known oxygen concentration: As For sample concentrations at or below ambient oxygen level (20.9%), expose the sensor to air. Adjust the "Calibrate" control until the meter reads 21. b. For sample concentrations above 20,9% and for best accuracy, expose the sensor to calibration gas with oxygen in a concentration higher than the sample (100% oxygen calibration gas is preferred). Use an adaptor to seal the sensor against outside air while exposing it to the sample or calibration gas and introduce a calibration gas. Adjust the "Calibrate" control to the actual content of the standard, whether you are using 100% oxygen or another certified calibration gas mixture. With the sensor in air, check the HI ALARM by setting the HI ALARM setpoint potentiometer below 21% oxygen, The audible and visual alarms should both activate. Readjust the HI ALARM setpoint to the desired level. Check the LO ALARM by setting its setpoint potentiometer above 21% oxygen. The audible and visual alarms should both activate. Readjust the LO ALARM setpoint to the desired level. Model 352 is ready for use, Model 353 1. 2. ds 4, Place the power switch in the ON position. If LO BAT appears on the display, then replace the batteries. Be sure that the B-5 electrochemical oxygen sensor has the sensor cable plugged into it. Connect the other end of the sensor cable into its receb- tacle on the front panel of the analyzer, Check the fixed low alarm by adjusting the "calibrate" control until the meter reads below 19.5%. The audible and visual alarms should both activate. Press the AUDIO SILENCE button and verify that the audible alarm silences for about 30 seconds. Calibrate the instrument to a known oxygen concentration: a. For sample concentrations at or below ambient oxygen level (20.9%), expose the sensor to air. Adjust the "Calibrate" control until the LCD reads 21. b. For sample concentrations above 20.9% and for best accuracy, expose the sensor to a calibration gas with oxygen in a concentration higher than the sample 11 6. (100% oxygen calibration gas is preferred). Use an adapter to seal the sensor against outside air while exposing it to the sample or calibration gas and introduce a calibration gas. Adjust the "Calibrate" control to the actual content of the standard, whether you are using 100% oxygen or another certified calibration gas mixture. | Model 353 is ready for use. 4.5 Model 354 1. Place the power switch in the ON position. If LO BAT appears on the display, then replace the batteries. Be sure that the B—5 electrochemical oxygen sensor has the sensor cable plugged into it. Connect the other end of the sensor cable into its гесер- tacle on the front panel of the analyzer. Calibrate the instrument to a known oxygen concentration: a. For sample concentrations at or below ambient oxygen level (20.9%), expose the sensor to air. Adjust the "Calibrate" control until the LCD reads 21. b. For sample concentrations above 20.9% and for best accuracy, expose the sensor to a calibration gas with oxygen ina concentration higher than the sample (100% oxygen calibration gas is preferred). Use an adapter to seal the sensor against outside air while exposing it to the sample or calibration gas and introduce a calibration gas. Adjust the "Calibrate" control to the actual content of the standard, whether you are using 100% oxygen or another certified level. With the sensor in air, check the HI ALARM by setting the digital HI ALARM setpoint potentiometer below 21% oxygen. The audible and visual alarms should both activate. Press the AUDIO SILENCE button and verify that the audible alarm silences for about 30 seconds. Readjust the HI ALARM setpoint to the desired level. Check the LO ALARM by setting the digital LO ALARM set- point potentiometer above 21% oxygen. The audible and visual alarms should both activate, Press the AUDIO SILENCE button and verify that the audible alarm si- lences for about 30 seconds. HReadjust the LO ALARM setpoint to the desired level. Model 354 is ready for use. 12 5 SERVICE INSTRUCTIONS 5.1 Troubleshooting The following is a general outline of the troubleshooting proce— dure, logic of the situation, and particular faults may require steps not mentioned. | 5. а. 1.1 .1.2 1.3 in practice, many steps could be skipped, according to the Preliminary Checks Check for physical damage. Be especially aware that damage due to abuse or misuse may void the warranty for the instrument or the Micro-Fuel Cell. Check the sensor: calibrate at 100% oxygen, then recheck in air; should read 21 +1% (Models 352, 353, and 354). Check the instrument with a good sensor. Check the sensor connections. Check the batteries, connections, and voltages. Measurement Circuit Checks Check reference voltages (if alarms are used). Check the sensor amplifier. Check the span pot continuity and operation. Alarm Circuit Checks (Models 351, 352, 353, and 354) Check the alarm pots for continuity and operation. Check the voltages at the inputs of the comparators. Check the comparator input/output match. If the positive input is of greater magnitude than the negative input, the output voltage should be positive; if the positive input is smaller than the negative input, then the output voltage should be negative. Check the voltage at each end of the setpoint potentiometer. Check the voltage at the slider to be sure that it corresponds with the physical setting. If all else checks okay, test the logic circuit (refer to Section 2.4.4 for more information). If you find no problems with the alarm circuit, and yet the LED does not light or the beeper does not beep, then test or replace the LED or beeper, as necessary. 13 5.2 5.1.4 Meter Circuit Checks a. Check the meter connections. b. Check the continuity between the main circuit and the meter. с. Check the meter response to changes in input signal. d. Check the word messages of digital meters by simulating the conditions that would activate them. For example, check the alarm messages by adjusting the setpoints until the instru- ment is in an alarm condition; check the "LO BAT" message by deliberately providing the instrument with insufficient battery voltage. Repair/Replacement CAUTION: ICs are susceptible to damage from static electricity. Ground yourself and your scldering iron before working on the unit. Wear natural fibers — no nylon or polyester! Do not touch the circuit boards with your fingers. Contamination from body salts or oils may affect the accuracy of the unit. Battery Requirements Model 350: None Models 351 and 352: Four Size "C" batteries Models 353 and 354: Two standard 9-volt batteries 5.2.1 Battery Replacement Models 351, 352, 353, and 354 1. Place the power switch in the OFF position. 2. Move the sliding back panel to the right to expose the battery compartment. 3. Remove old batteries. 4. Install the correct batteries, observing proper polarity. 5, Close the sliding back panel. 14 5.3 5.2.2 Sensor Replacement CAUTION: Do not scratch, puncture, or otherwise damage the sensor's mem- brane. Damage to the membrane will require sensor replacement. NEVER PRESS ON THE SENSING SURFACE: you might damage the sensor. Micro-Fuel Cell - B-6 1. Remove the new B-6 Micro-Fuel Cell from its protective bag. Carefully remove the shorting clip. 2.. Unscrew the holder cap from the sensor holder and remove the existing Micro-Fuel Cell, 3. Place the new Micro-Fuel Cell inside the sensor holder with the cell's membrane surface facing outward, and the electri- cal contacts facing inward. 4. Screw the cap back onto the sensor holder. 5. Check to see that the sensor cable is plugged into the receptacle on the front of the analyzer. Micro-Fuel Cell — B-5 i. Remove the new B-5 Micro-Fuel cell assembly from its protective bag. 2. Unplug the coiled cord from the old B-5 cell assembly. 3. Plug the coiled cord into the new B—5 cell assembly. a, Discard the old B-5 cell assembly. 5.2.3 Disassembly ~ All Models Hemove the four screws at the corners of the underside of the instrument, then pull the top cover up and away from the electronics front panel and PC board. Manufacturing Test Procedures The following test procedures are those actually used in Tele- dyne’s testing department. They may provide information or sug- gestions on how to perform troubleshooting tests, as well, so we are including them with your service manual. Please note that the test procedures depend upon the use of a "simulator" to take the place of the measuring cell during test- ing of the instrument. The cell simulator can be purchased from Teledyne Electronic Devices. 15 1.5 Y Battery: Or, a simple circuit such as the one illustrated in the sketch can be constructed to substitute for the simulator. Micro-Fuel Cell Simulator 5 K AAA) 1.5 volt battery Current Source 150 K 10 K + 0 T Adjust for 21 mv O Voltage Source 16 TEST PROCEDURE MODEL 350 Be sure that the meter reads zero: adjust the mechanical meter zero, if necessary. Connect the probe assembly. Set the Micro-Fuel Cell simulator selector switch to the C-1 position: connect the simulator to the probe assembly (positive lead to the outer spring contact, negative lead to the center spring contact), Simulate 100% oxygen and adjust the CALIBRATE control until the meter reads 100. Similate 20.9% oxygen. The meter needle should coincide with the red CAL mark on the meter scale. Read just the mechanical meter zero until the needle points exactly to the CAL mark. 17 10. il. TEST PROCEDURE MODELS 351 and 352 Before supplying power to the unit, be certain that the pointer of the readout meter points exactly to zero. Adjust the mechan- ical zero of the meter, if necessary, Slide open the battery panel at the rear of the instrument, and install four size "C" alkaline batteries, observing proper polar- ity as shown. Place the power switch in the ON position. Adjust trimmer P-1 on the PC Board until the meter reads exactly zero. Connect the probe assembly. Set the Micro-Fuel Cell simulator to a 70 microampere output with the selector switch in the A-3 pos- ition and the attenuator in the CAL position. Connect the simul- ator to the probe assembly (positive lead to the outer spring contact and the negative lead to the center spring contact). Set the attenuator of the simulator to 100 %. Adjust the CALI- BRATE control of the test unit until the meter reads 100. Set the attenuator of the simulator to CAL (21%). The meter needle should coincide with the red CAL mark on the meter scale, Readjust the mechanical meter zero until the needle points exact- ly to the CAL mark, Set the HIGH limit to read 30% and the LOW limit to read 25%. Offset the CALIBRATE control so that the display reads above 30%. The HIGH warning (HI ALARM) (Model 352 only) should be shown on the display and the light should flash. The audible alarm should sound, as well. Alarming should occur within + digit of the HI ALARM setpoint. Turn the CALIBRATE control until the meter reads below 25%; LOW warning should appear on the meter within +1 digit of the set- point, and the audible and visual alarms should energize. Set the LOW limit to 10%. Slowly adjust the CALIBRATE control to lower the meter reading; the LOW warning should appear at 19.5% +2%. The audible and visual alarms should energize. Turn the unit off and replace the two batteries in the positive half of the power supply with a variable DC power supply. Place a capacitor (300 to 600 mfd.) across the power supply to lower the source impedance. Turn the unit on and vary the supply voltage from 3 volts to 1.8 volts while holding the power switch in the BAT TEST position. The pointer of the meter should cross the BATTERY REPLACE mark when the supply voltage is 2.0 +0.1 volts. 18 10. il. TEST PHOCEDURE MODELS 353 and 354 Connect a 9V battery to the battery terminals, and switch on the instrument. Connect a current meter across the power switch terminals. Switch off the instrument. The current reading should be below 1 má. Switch the instrument on again and remove the current meter. Connect the DVM negative lead to pt. #5 (ground) on the main PC Board and the positive lead to pt. #7. The DVM should indicate +4.5 V, or half the battery voltage. Connect DVM positive to pt. #9. DVM should read -4.5 V, or half the battery voltage. “Connect DVM positive lead to pt. #2; DVM should read 2.00 +.02 V. Adjust trimmer P-1 on the PC Board until the meter reads exactly Zero. Connect the probe assembly. Set the Micro-Fuel Cell simulator to a 70 microampere output with the selector switch in the 4-3 position and the attenuator in the CAL position. Connect the simulator to the probe assembly (positive lead to the outer spring contact and the negative lead to the center spring contact}. Set the attenuator of the simulator to 100 %. Adjust the CALI- BRATE control of the test unit so that the voltage out of A3a at pin 16 is 2.000 volts. The display should read 100 +1. Check the "SIG OUT" voltage; it should be 100 +2 millivolts. Set the attenuator of the simulator to CAL (21%). The display should read 21 +1. Set the HIGH limit to read 30% and the LOW limit to read 25%. Offset the CALIBRATE control so that the display reads above 30%. The HIGH warning (HI ALARM) (Model 354 only) should be shown on the display and the light should flash. The audible alarm should sound also. Press the silence pushbutton; the audible alarm should silence for 30 +10 seconds. Alarming should occur within + digit of the HI ALARM setpoint. Turn the CALIBRATE control until the meter reads below 25%; LOW warning should appear on the meter within +1 digit of the set- point, and the audible and visual alarms should energize. 19 12. 13. Set the LOW limit to 10%. Slowly adjust the CALIBRATE control to lower the meter reading; the LOW warning should appear at 19.5% +1%. The audible and visual alarms should energize. Disconnect the battery and connect a variable power supply bet- ween the battery leads. Vary the supply voltage from Y volts to 6.5 volts. At 6.5 +.2 volts, the LO BAT indicator should appear. 20 SPECIFICATIONS Model Number 350 351 352 353 354 0-25% Range YES YES NO NO NO 0-100% Range NO NO YES YES YES Analos Meter YES YES YES NO NO Digital Meter (2 1/2 digit LCD) NO NO NO YES YES Accuracy at Constant Temperature +2% +2% +2% +1% +1% (% of Fuliscale) Accuracy over Temperature Range +5% +5% +5% +5% +5% (% of reading) 90% Response Time (Seconds) 13 13 13 13 13 0-100mV Voltage Output NO NO NO YES YES Fixed Low Alarm at 19.5% 02 NO YES NO YES NO 2 Adjustable Alarms, (HI, LOW) NO NO YES NO YES Audible and visual alarms NO YES YES YES YES Temporary Audible Alarm Silence NO NO NO YES YES MicroFuel Cell Type B~6 B-5 B-5 B-5 B-5 Warrantv/Expected | Life (Months in Air) 18/36 18/36 18/36 18/36 18/36 Battery Type None 4-0 4-C 2-9v 2-9v Battery Expected Life In Months for Alkaline Cell - 8 8 8 8 Height, inches (mm) = ARA 5.8(147)H--————————————— Width, inches (mm) .6(218) 5(267) 10.5(267) 8.6(218) 6(218) Depth, inches (mm) .0(76) 5(114) 4,5(114) 4.5(114) 5(114) Weight, in pounds 1.9 3.2 3.3 2.5 2.6 Weight, in grams (860) (1400) (1500) (1100) 1200) Probe Length (feet) 8 8 8 8 8 Operating Temperature Range a 0-500C/32~122°F LJ EE eo er A Pr a CELL MICRO-FUEL CELLS FOR MODELS 350, 551, 352, 353, AND 354 WARRANTED PEHCENT OXYGEN ANALYZERS EXPECTED 90% QUTPUT CLASS LIFE LIFE HESPONSE IN AIR B—5 18 MOS. 36 MOS. 13 SEC. 21mVY +20% IN AIR B-6 18 MOS. 36 MOS. 13 SEC. .30mA +20% IN AIR NOTE: Specifications are at 25°C (77°F), a MATERIAL SAFETY DATA SHEET Section I ~ Product Identification Product Name: Micro Fuel Cells r all classes execpt A~2, A-3, | А-5, B-2F and B-2CF. Electrochemical Oxygen Sensors, all classes execpt R-19, Mini micro-fuel cells, all classes, Manufacturer: Teledyne Analytical Instruments Address: 16839 Chestnut Street, City of Industry, CA 91749 Phone Number: (818) 961-9221 MSDS Prepared By: Chuck Molloy Date Prepared or Last Revised: 3-31-88 Emergency Phone No.: (818) 961-9221 section IT ~ Physical and Chemical Data — Chemical and Common Names: Potassium Hydroxide (KOH) 15% Granular Lead (PP) pure CAS Number: KOH 1319-58-32 Pb 7439-921 Melting Point/Range: -10 to 0 Deg. C Boiling Point/Range: 108 to 115 Ред. С Specific Gravity: 1.09 @ 20 Ред. С pH: >14 Solubility in Waters: Completely soluble Percent Volatiles by Volume: None Appearance and Odor: Colorless, odorless solution li Section III - Physical Hazards Potential for Fire and Explosion: The electrolyte in micro-fuel Cells is not flammable. There are no fire or explosion hazards associated with micro-fuel cells. Potential for Reactivity: The sensors are stable under normal conditions of use. Avoid contact between the sensor electrolyte and strong acids. ei ii Tr rushers mala Primary Route(s) of Entry: Ingestion, Eye/Skin Contact Exposure Limits: OSHA PEL .05 mg/cu.m. (Pb) ACGIH TLV 2 mg/cu.m. (KOH) Effects of Overexposure: Ingestion: The electrolyte could be harmful or fatal if swallowed. Oral 1050 (ВАТ) = 2433 mg/kg Eye: The electrolyte is corrosive and eye contact could result in permanent loss of vision. Dermal: The electrolyte is corrosive and skin contact could result in a chemical burn. | Inhalation: Liquid inhalation is unlikely. Signs/Symptoms of Exposure: Contact with skin or eyes will cause a burning sensation and/or a soapy feeling for skin contact. Medical Conditions Aggravated by Exposure: None Carcinogenicity: NIP Annual Report on Carcinogens: Not listed. LARC Monographs: Not listed. OSHA: Not listed. Other Health Hazards: Lead is listed as a chemical known to the State of California to cause birth defects or other reproductive harm. Section Y - Emergency and First Aid Procedures Eye Contact: Flush eyes with water for at least 15 minutes and get immediate medical attention. | ` Tr aT REL CELLS, MSD Skin Contact: Wash affected area with plenty of water and remove contaminated clothing, If burning persists, seek . medical attention. RE Ingestion: Give plenty of cold water. Do not induce vomiting, Co Get medical attention. Lo e Inhalation: Liquid inhalation is unlikely, Section VI - andling Information Handling Precautions: Note: The Oxygen sensors are sealed and under normal Circumstances, the contents of the sensors do not present a health hazard. The following information is given as a guide in the event that a cell leaks. Protective Clothing: Rubber gloves, chemical splash goggles. Clean-up Procedures: Wipe down the area several times with a wet paper towel. Use a fresh towel each time. Protective Measures during Cell Replacement: Before opening the bag containing the cell, check the cell for leakage. If the cell leaks, do not open the bag. If there is liquid around the cell while in the instrument, put on gloves and eye protection before removing the cell. Disposal: Should be in accordance with all applicable state, local and federal regulations. ea CF Гу 4, . Da Ta Le - _ * pram as ; > SE SPARE PARTS LIST FOR MODEL 354 QIY. PART NUMBER DESCRIPTION Al76 ALARM Ese | CLIP AND WIRE, BATTERY B30171 ASSEMBLY, PC BOARD (ICD) B40304 ASSEMBLY, PC BOARD B39955 — SENSOR, CLASS B-5 1 K35 KNOB 1 Jl JACK 1 L154 LED 1 L156 LENS, RED 1 P322 POT, 5 K (Bourns #3852-282-502A) |" * 2 P443 POT, 100 K (DIGITAL) A MINIMUM CHARGE OF $20.00 I5 APPLICABLE TO SPARE PARTS ORDERS. IMPORTANT: Orders for replacement parts must include the model and serial number of the equipment for which the parts are intended. This information is available on the Specific Application page in the front of this manual. Send orders to: TELEDYNE ANALYTICAL INSTRUMENTS 16830 CHESTNUT STREET CITY OF INDUSTRY, CA 91740-1580 TELEPHONE: (818) 961-9221 TWX: 910-584-1887 TDYANYL COID REFERENCE DRAWINGS MODEL 354 C~40303 OUTLINE DIAGRAM C-40306 SCHEMATIC C-40308 WIRING DIAGRAM ee an a Te] We MIZATYNY NIDAXO : DSE TON ыы В Ен Бой gro fv An 5 CWI AgISOONI AO AD Te eE ur .. ej SiNZANELSNIIVOMATNY poa (65 Fie то Y == | | ji | zoo SNAQ=13124 t= = лаб ROISSINEDe NEL ITA LOMA CIEN HO CINCA GEO JE CL LON SI 16 ROM CNTUOINE TT LENSOR GT Ce te - Mae, Ch WA D ONY IRA OI131 30 AFETee FAL SE DAA RIAD TIM ENE 1 AOL S30 | "ово FE | re LT J | i : 3 | | 2 социо puise ql Jue dy rp - CSStISNEQ HOR E DUMPYIE 10 Éxtasis ains A yee JOUE Poe SOURED {но LAO | PANES CHIP OT PULPO Cry Gio 7 - AJO AS mod vn) ego vo Saar ¡78 015 4 SHEZIiIVvO IYIdIg OL ONE Giro ARE pon pa O YO! an waldo seed ay 11607) 30 Nut Tuy faeudas Arpintepro Y nur ‘нада! RO SA Ap sp 1 CZ “MITE remibs uE срыв Арины Эрнан Had porra y 4 TIVHENOWSD 02 py ny EN Eve EM N IMHO HE FI BEE HE CE NE BCE LET SHINY SKE 2 5 e 8 * y Me x Е Эх 4 sv .XOR 10 NOISNIMIO NI? WR y 2 SHUIITHITHIW YO Sil Y SIHONL NI JYY SNOISNIWIG TY ("| S IHINI | - 210N 4 í - Y || Нее нанят те meee nm aan сен с тут пе бе ее mee me SEP | Is TIA SiINZIWAHMLSMI TYOLLATYNW v- y NS | A INAGITII LIM, o ASSy COIPE-E САО L2%OVEB ONLINNOW 87. b eS GIQVIEHL O7-Y/71 G О у Cir] Si” 4 O € Y - \ (ВОР) че y “с 5 , ad poros E Pp Y a 848 | a i зв 553 Ё SMÉTRIAIN | YO | ak NIZE | ть i Hsty CH UNIS ASTRAL 1 + 90505 5 CIYLIDIC)? DIAYWIEHDS 0.46 es Sb Dj dir EE Y NOUIADN YA зе OSO Yo Bl cet waa] Nt wo wa том NASAXLO +55 TOWN mn HER a FPLIA VINO 1D ANLSOONI ID ALIS ти EVITA Nive SINGSHNEISNE TYOLLAIVNY 55 > DEA 16 | ~~ ars ENAGST LEE Fea WE NOISSIMHIE Linn ; , - SH QUO 36 OL LON Si LI HON VANO NI TWiINEO EA minos a 03005 NADIE JO AIBIGOUS 3H1 $1 ONIMVUO SH RANTE NOtLdIHDSZ0 | - a2 nd [wan 1 A o ST 1 Crs) ADI 1 AOL 2 ' MS i PA OWA + = sh +1 1® сом) MNOLLNG Mod | 25 se A, O. 6 TA = | ao WAS © AAAS L a er AS%- A i wol PA i Sneed } ANA $ sa i «| Зы Se } marre — _ } nam = AB : | “= fam” SEE ra x | + а 3 ! iva] At js я a "1 = al a BLE ! > “oo | visa) | hel $ess | | 3 ч al 1 == > mo q 1 мо < ен ) ВУЗ $607 a we Te y > 1] Р AAA mm ere rere ee {ps mmm = Bla EEE у ! уни day AZ e Gy 4 wot VAN GC; vie Au aha © + de A A+ Вы! + ex IAS DL “On > “Ed AEZI- A A MA SOEWI “Oo ora ul AO SEH = la | ANO o э ор Va < AAA GA ¢ || fom] steuer ay Wo gas eed po AED Sb vis - 1 ea oe | St SEIN 4001 as a AAAO PA > Les Wz a A GO 1 < g AAA ue o AAN & a A C— сн eL ! € or Bib #5 << ё Bi Ko : ey *or ir ARE ie BARS HOLM Zn so Le 4 po Vie E dos € | 92% EAS “PENE er АС = 7-16 = OTO-G6 HOIT (ET SIL! CS a SU Ib 440731 рые кН | Y SU rr 95% i ly 2 (07-Y9g ‘91—a¥) ¥3d338 WONJ SIJIM YE NY Ce ASSY 83d INN Sd (£ 310N 335) HOLIINNOO AMBLIVE № de TA SILIN vilSia &did1J6 @ CRS ON |B] ON IB 215 NE OY TE—HA L8/97/3 9160-26 #021 "ON 15/5171 0400-156 #003 "ON "day ЗАЗ NOAA SNOISIAJE y 200% E —— | +-восово Gl do | zonmmau or aug | ds os zoo [4 HONG /0 #55 10N == 2 26/tt/€ СУОЗНал М | Эно ИМ WYHOVIA ONISIA 28/0178 SEA jo ñ Tun] aw SHOES 9050970 /5 ELLE VIDIDI7TVO AMISGON) 40 AO Fa saIborugoa| ooleg etápojo: JO YUN ESCUIENG y = = AXE UVANT SIUSLUNIFSUI JPONADUY ouAPSIEL Ja ZF nowy saunas 94G FIVOS LON 06 ON Ja | Ash | Mas MAA WOLLOE (N/d HOIDANNOD AMILIVE HOJ 1Z/7£-8 'OMO 338} ‘007 JOlAMIS MOTIY OL HONONI 9NOT SEYIM IS ONY 18 — DIVA “SIRALIYO OL T8YO, NUOS OL SEll lods daY (5) (SId ©) СУММЫ) 3NO OL 038307108 o SIUIM OML COIY (4) HIM CIMNVA SNOLIINNOS Э VAAYE—H AVHOVIG ONISIM TINYd LNOS3 WO) DNILSXI 3ev SANTI CALLOO NI NAOHS ONISIM ION TINYd ONY Bd NIIMII9 SINIM “IMY ZZ 38 OL SIMIM o G314103dS ISIMUJHLO SSIINN SALON © y
advertisement
Related manuals
advertisement