System Verification and
Performance Test Guide
HP 8753E Network Analyzer
7) Pacino
HP Part No. 08753-90394 Supersedes January 1998
Printed in USA January 15, 1998
Notice.
——
The information contained in this document is subject to change without notice. I
Hewlett-Packard makes no warranty of any kind with regard to this material, including
but not limited to, the implied warranties of merchantability and fitness for a particular
purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or —
consequential damages in connection with the furnishing, performance, or use of this material.
“a” —
O Copyright Hewlett-Packard Company 1998
All Rights Reserved. Reproduction, adaptation, or translation without prior written permission
is prohibited, except as allowed under the copyright laws.
1400 Fountaingrove Parkway, Santa Rosa, CA 95403-1799, USA
Certification
Hewlett-Packard Company certifies that this product met its published specifications at the
time of shipment from the factory. Hewlett-Packard further certifies that its calibration
measurements are traceable to the United States National Institute of Standards and
Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration
facilities of other International Standards Organization members.
Warranty
This Hewlett-Packard instrument product is warranted against defects in material and
workmanship for a period of three years from date of shipment. During the warranty period,
Hewlett-Packard Company will, at its option, either repair or replace products which prove to
be defective.
For warranty service or repair, this product must be returned to a service facility designated by
Hewlett-Packard. Buyer shall prepay shipping charges to Hewlett-Packard and Hewlett-Packard
shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all
shipping charges, duties, and taxes for products returned to Hewlett-Packard from another
country.
Hewlett-Packard warrants that its software and firmware designated by Hewlett-Packard for
use with an instrument will execute its programming instructions when properly installed on
that instrument. Hewlett-Packard does not warrant that the operation of the instrument, or
software, or firmware will be uninterrupted or error-free.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate
maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or
misuse, operation outside of the environmental specifications for the product, or improper
site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. HEWLETT-PACKARD SPECIFICALLY
DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
HEWLETT-PACKARD SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL,
INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT,
OR ANY OTHER LEGAL THEORY.
Assistance
Product maintenance agreements and other customer assistance agreements are available for
Hewlett-Packard products. For any assistance, contact your nearest Hewlett-Packard Sales and
Service Office.
Hewlett-Packard Sales and Service Offices
US FIELD OPERATIONS
Headquarters
Hewlett-Packard Co.
19320 Pruneridge Avenue
Cupertino, CA 95014
(800) 752-0900
Colorado
Hewlett-Packard Co.
24 Inverness Place, East
Englewood, CO 80112
(303) 649-5512
New Jersey
Hewlett-Packard Co.
150 Green Pond Rd.
Rockaway, NJ 07866
California, Northern
Hewlett-Packard Co.
301 E. Evelyn
Mountain View, CA 94041
(415) 694-2000
Atlanta Annex
Hewlett-Packard Co.
2124 Barrett Park Drive
Kennesaw, GA 30144
(404) 648-0000
Texas
Hewlett-Packard Co.
930 E. Campbell Rd.
Richardson, TX 75081
California, Southern
Hewlett-Packard Co.
1421 South Manhattan Ave.
Fullerton, CA 92631
(714) 999-6700
Illinois
Hewlett-Packard Co.
545 E. Algonquin Rd.
Arlington Heights, IL 60005
(847) 342-2000
(201) 586-5400 (214) 231-6101
EUROPEAN FIELD OPERATIONS
Headquarters France Germany
Hewlett-Packard S.A. Hewlett-Packard France Hewlett-Packard GmbH
150, Route du Nant-d’ Avril
1217 Meyrin 2/Geneva
Switzerland
(41 22) 780.8111
Great Britain
Hewlett-Packard Ltd.
Eskdale Road, Winnersh Triangle
Wokingham, Berkshire RG41 5DZ
England
(44 734) 696622
1 Avenue Du Canada
Zone D'Activite De Courtaboeuf
F-91947 Les Ulis Cedex
France
(33 1) 69 82 60 60
Hewlett-Packard Strasse
61352 Bad Homburg v.d.H
Germany
(49 6172) 16-0
INTERCON FIELD OPERATIONS
Headquarters
Hewlett-Packard Company
3495 Deer Creek Road
Palo Alto, California, USA
94304-1316
(415) 857-5027
China
China Hewlett-Packard Company
38 Bei San Huan X1 Road
Shuang Yu Shu
Hai Dian District
Beijing, China
(86 1) 256-6888
Taiwan
Hewlett-Packard Taiwan
| 8th Floor, H-P Building
337 Fu Hsing North Road
Taipei, Taiwan
(886 2) 712-0404
Australia
Hewlett-Packard Australia Ltd.
31-41 Joseph Street
Blackburn, Victoria 3130
(61 3) 895-2895
Japan
Hewlett-Packard Japan, Ltd.
9-1 Takakura-Cho, Hachioji
Tokyo 192, Japan
(81 426) 60-2111
Canada
Hewlett-Packard (Canada) Ltd.
17500 South Service Road
Trans-Canada Highway
Kirkland, Quebec H9J 2X8
Canada
(514) 697-4232
Singapore
Hewlett-Packard Singapore (Pte.) Ltd.
150 Beach Road
1429-00 Gateway West
Singapore 0718
(65) 291-9088
Safety Notes
The following safety notes are used throughout this manual. Familiarize yourself with each of
the notes and its meaning before operating this instrument.
Caution
Caution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, would result in damage to or destruction
of the instrument. Do not proceed beyond a caution sign until the indicated
conditions are fully understood and met.
Caution
Always use the three-prong ac power cord supplied with this instrument.
Failure to ensure adequate earth grounding by not using this cord may cause
instrument damage.
Warning
Warning denotes a hazard. It calls attention to a procedure which, if not
correctly performed or adhered to, could result in injury or loss of life.
Do not proceed beyond a warning note until the indicated conditions are
fully understood and met.
Warning
No operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock, do not remove covers.
warning
For continued protection against fire hazard replace line fuse only with
same type and rating (F 3A/250V). The use of other fuses or material is
prohibited.
Warning
This is a Safety Class I product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be
inserted in a socket outlet provided with a protective earth contact. Any
interruption of the protective conductor, inside or outside the instrument,
is likely to make the instrument dangerous. Intentional interruption is
prohibited.
Warning
The power cord is connected to internal capacitors that may remain live
for 10 seconds after disconnecting the plug from its power supply.
warning
The opening of covers or removal of parts is likely to expose dangerous
voltages. Disconnect the instrument from all voltage sources while it is
being opened.
The instruction documentation symbol. The product is marked with this symbol when it
is necessary for the user to refer to the instructions in the documentation.
“CE” The CE mark is a registered trademark of the European Community. (If accompanied by
a year, it is when the design was proven.)
“ISMI-A” This is a symbol of an Industrial Scientific and Medical Group 1 Class A product.
“CSA” The CSA mark is a registered trademark of the Canadian Standards Association.
HP 8753E Network Analyzer Documentation Set
vi
The Installation and Quick Start Guide
familiarizes you with the HP 8753E network
analyzer’s front and rear panels, electrical and
environmental operating requirements, as well
as procedures for installing, configuring, and
verifying the operation of the HP 8753E.
The User's Guide shows how to make
measurements, explains commonly-used
features, and tells you how to get the most
performance from your analyzer.
The Quick Reference Guide provides a
summary of selected user features.
The HP-IB Programming and Command
Reference Guide provides programming
information for operation of the network
analyzer under HP-IB control.
The HP BASIC Programming Examples
Guide provides a tutorial introduction using
BASIC programming examples to demonstrate
the remote operation of the network analyzer.
The System Verification and Test Guide
provides the system verification and
performance tests and the Performance Test
Record for your analyzer.
Contents
1.
System Verification and Performance Tests
How to Test the Performance of Your Analyzer
Sections in this Chapter . . . . . . . aaa aaa
Performance Test Record . . . . . . . . . . . .
- * *
= os Fo» x x ow a ıxı=*
чай 0=
System Verification Cycle and Kit Re-certification . . . . . . . . . . . . . ..
HP 8753E System Verification . . . . . . . . .
Ex 0&8 & & a
Initialization . . . . . . . . . . . Cee ee ee ee
Device Verification . . . . . . . . . . . . . . . .. Coe.
In Case of Difficulty . . . . . . . . 2 1 11 LL LL LL LL LL 111110
1. Test Port Output Frequency Range and Accuracy
In Case of Difficulty . . . . . . . . . . .... .
E 4 - *
2. External Source Mode Frequency Range . . . . . . . ..
In Case of Difficulty . . . . . . . . . . . ..
# % #0 #1 #1 ww
w a #8 #8 вв
BEAR x» "w=
EE EE IE A
#1 #1 #8 ER = ME = 0»
oo #0 A
a * LE O A A EE
3. Test Port Output Power Accuracy . . . . . . . . . . . . . . . .. . . ..
In Case of Difficulty . . . . . . . . . . . aaa
4. Test Port Output Power Range and Linearity
In Case of Difficulty . . . . . . . . . aaa
“ С >
5. Minimum R Channel Level. . . . . . . . . .. CLL.
In Case of Difficulty . . . . . . .. . . ..... aaa
6. Test Port Input Noise Floor Level . . . . . . . . . . . . . . . . . ....
Port 1 Noise Floor Level from 300 kHz to 3GHz (IF BW = 3 kHz) . . . . . . .
Port 1 Noise Floor Level from 300 kHz to 3GHz(IFBW = 10Hz) . . . . . . .
Port 2 Noise Floor Level from 300 kHz to 3 GHz (IF BW =
EE
* #8 A Wa
OE E E RN
Port 2 Noise Floor Level from 300 kHz to 3 GHz (IF BW = 3kHz). .. . . . .
BW =3kHz). . .. .. .. |
Port 2 Noise Floor Level from 3 GHz to 6 GHz (IF
Port 2 Noise Floor Level from 3 GHz to 6 GHz (IF
BW
|
1I0Hz) . . . . . ...
Port 1 Noise Floor Level for 3 GHz to 6 GHz (IF BW = 10 Hz)
LA каж 8
Port 1 Noise Floor Level from 3 GHz to 6 GHz (IF BW = 3kHz). .. . . . ..
In Case of Difficulty . . . 2. 22 112 2 4 LL LVL LL LL 1111 LS
7. Test Port Input Frequency Response . . . . . . . . . ..
Power Meter Calibration for Test Port 1 from 300 kHz to 3 GHZ . . . . . . . .
Test Port 2 Input Frequency Response from 300 kHz to 3 GHz
Power Meter Calibration on Port 2 from 300 kHz to 3 GHz
Test Port 1 Input Frequency Response from 300 kHz to 3 GHz
Power Meter Calibration for Test Port 2 from 3 GHz to 6 GHz
Test Port 1 Input Frequency Response from 3 GHz to 6 GHz
Power Meter Calibration on Test Port 1 from 3 GHz to 6 GHz .
Test Port 2 Input Frequency Response from 3 GHz to 6 GHz
In Case of Difficulty . . . . . . . . . . .. ... . ...
8. Test Port Crosstalk . . . . . . . . . . . ..
~ Crosstalk to Test Port 2 from 300 kHz to 3 GHz
Crosstalk to Test Port 1 from 300 kHz to 3 GHz
тока вв
- * “ +
# 250% PD O 4
« 4a 5 84 & = & =
0 a Fw #8 = =
№ 5х
ESE E ER 0%
ON A EE 0k
токов
® B® 0» 0» a = x >
® A вова
OE EC FH = oa W
¥ Fs * A PP ®*
OE OB RB x ® ow
Crosstalk to Test Port 1 from 3 GHzto6 GHz . . . . . . . . . . . . . . ..
Crosstalk to Test Port 2 from 3 GHzto 6 GHz . . . . . . . . . . . . . . ..
In Case of Difficulty . . . . . . . . . . ......
9. Calibration Coefficients .
First Full 2-Port Calibration .. 2.1 2 2214 4 4 1 LL 4 LL 11 1210
Directivity (Forward) Calibration Coefficient LL 4 4 41114 4 120
Source Match (Forward) Calibration Coefficient . . . . . . . aa
Transmission Tracking (Forward) Calibration Coefficient . . . . . . . Co.
Reflection Tracking (Forward) Calibration Coefficient . . . . . . . . . . . . .
Load Match (Reverse) Calibration Coefficient . . . . . . . . . . . . . ...
Transmission Tracking (Reverse) Calibration Coefficient . . . . . . . . . . .
Second Full 2-Port Calibration . . . . . . ............. Coe.
Load Match (Forward) Calibration Coefficient . . . . . . . . . . . . . . ..
Directivity (Reverse) Calibration Coefficient Ce ee aaa a aa
EEE EEE BE A E AE EA EA
Reflection Tracking (Reverse) Calibration Coefficient . . . . . . . . .
10. System Trace Noise (Only for Analyzers without Option 006) . 2.
System Trace Noise for A/R Magnitude . . . . . . . . . . 2... .....
System Trace Noise for A/R Phase . . . . . . . . . . .. a ea a aa
System Trace Noise for B/R Magnitude . . . . . . . . 2 ene. |
System Trace Noise for B/R Phase . . . . . . . . . . . . e ee.
In Case of Difficulty . . . . . . . . ©. 4 LL LL LL LL LL 1110
11. System Trace Noise (Only for Analyzers with Option 006) . . . . . . . . .
System Trace Noise for A/R Magnitude from 30 kHzto 3 GHz . . . . . . . . .
System Trace Noise for A/R Magnitude from 3GHzto6 GHz . . . . . . . . .
System Trace Noise for A/R Phase from 3 GHzto6 GHz . . . . . . . . . ..
System Trace Noise for A/R Phase from 30 kHzto3GHz . . . . . . . . . ..
System Trace Noise for B/R Magnitude from 30 kHzto 3 GHz . . . . . . . .
System Trace Noise for B/R Magnitude from 3 GHzto6 GHz . . . . . . . . .
System Trace Noise for B/R Phase from 3 GHzto6 GHz . . . . . . . . . ..
System Trace Noise for B/R Phase from 30 kHzto3 GHz . . . . . . . . 2...
In Case of Difficulty . . . . 2 1 21 22141 LV LL LL LL 1111220
12. Test Port Input Impedance .. . . 2. 2.124444 41111110
In Case of Difficulty . . . . 2 1 11 1 1 LL LL LL LL 1 11110 Coe.
13. Test Port Receiver Magnitude Dynamic Accuracy . . . . . . . . . . . . ..
Initial Calculations . . . . . aaa e ee a a a aa aa aa
Port 1 Power Meter Calibration aaa e aaa -
Adapter Removal Calibration . . . . . 2122 14 4 4 1 1 LL 111110
Measure Test Port 2 Magnitude Dynamic Accuracy . . . . . . . . . . . . ..
Measure Test Port 1 Magnitude Dynamic Accuracy . . . . . . . . . . . . ..
In Case of Difficulty . . . . 2122 21 4 4 LL LL LL LL 11110
14. Test Port Receiver Magnitude Compression . . . . . . . . . . . . . . ..
Test Port 2 Magnitude Compression . . . . . . . .. .. .. a e aa
Test Port 1 Magnitude Compression . . . . . . Ce ee ee ee ee ee,
In Case of Difficulty . . . . 2 221 2 4 41 1 4 1 1 LL 1110 Coe
15. Test Port Receiver Phase Compression . . . . . . . . . . . . .. . . .. .
Test Port 2 Phase Compression . . . . . . . . . . . . . . . . . . . ....
Test Port 1 Phase Compression . . . . . . . . . . aa aa an
In Case of Difficulty . . . . . . . . . ue ee
16. Test Port Output/Input Harmonics (Option 002 Analyzers without Option 006
Only) 2. 22 21 1 4 4244 44 4 EC 4 4 4 4 4 AL LL 1 120
Test Port Output Worst Case 2nd Harmonic .. . .. 1.141141 11140
Test Port Output Worst Case 3rd Harmonic . . . . . . . . . . . . . . ...
Port 1 Input Worst Case 2nd Harmonic . . . . . . . . . . .. a aa
Port 1 Input Worst Case 3rd Harmonic . . . . . . . . . . . . . .. . . ...
Port 2 Input Worst Case 2nd Harmonic . . . . . . . . . aaa ea
Port 2 Input Worst Case 3rd Harmonic . . . . . . eee
17. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 Only)
45
45
47
47
47
47
47
47
48
49
49
49
49
50
51
51
ol
51
51
52
53
53
53
53
53
54
54
54
54
55
58
59
61
62
64
66
68
68
69
69
70
71
72
72
73
73
74
74
75
76
76
TT
Ti
78
Test Port Output Worst Case 2nd Harmonic . . . . . . . . . . . .
Test Port Output Worst Case 3rd Harmonic . . . . . . . . . .. .
Port 1 Input Worst Case 2nd Harmonic . . . . . . . . . . . . ..
Port 1 Input Worst Case 3rd Harmonic . . . . . . . . . . . . . ..
Port 2 Input Worst Case 2nd Harmonic . . . . . . . . . . . . . .
Port 2 Input Worst Case 3rd Harmonic . . .
2. Performance Test Record
For Analyzers with a Frequency Range of 30 kHz to 3 GHz . . . .
3. Performance Test Record
For Analyzers with a Frequency Range of 30 kHz to 6 GHz . . . . . .
Index
dE EE #
EE EE
EE EE
EE EA
У ак
EA
* a =“ * *
* 0® 5% = +
79
80
81
82
82
82
Figures
_
1-1, System Verification Test Setup . . . . . 1 21 21 11112 10101110 5
1-2. Connections for Measurement Calibration Standards a aa aan 6
1-3. Transmission Calibration Setup . . . . ее. a aaa 7
1-4. Connections for the 20 dB Verification Device ae aaa e aa 8
1-5. Connections for the 50 dB Verification Device . . . . . . . . . . . . .. 8
1-6. Mismatch Device Verification Setup . . . . . . . . . . . . . 111110 9
1-7. Mismatch Device Verification Setup . . . . . . . . . . . . . . . . . .. 9
1-8. Test Port Output Frequency Range and Accuracy Test Setup . . . . . . . 12
1-9. External Source Mode Frequency Range Test Setup . . . . . . . . . . . 13
1-10. Source Output Power Accuracy Test Setup . . . . . . . . . . . . . ... 16
1-11. Test Port Output Power Range and Accuracy Test Setup . . . . . . . . . 19
1-12. Minimum R Channel Level Test Setup . . . . . . . . . . . . . . . .... 21
1-13. Flexible RF Cable Location . . . . . . . . . . . . . . . . . . ..... 23
1-14. Connections for Substituting the R Sampler (A4) . Ве, 24
1-15. Setup for Checking the R Sampler (A4) . . . . . . . . . . . . . .... 24
1-16. Source Input Noise Floor Test Setup . . . . . . . . . . . . . . . .... 27
1-17. Setup for Power Meter Calibrationon Test Port1 . . . . . . . . . . . . 31
1-18. Test Port 2 Input Frequency Response Test Setup . . . . . . . . . . . . 34
1-19. Setup for Power Meter Calibration on Test Port 2 . . . . . . . . . . . . 35
1-20. Test Port 1 Input Frequency Response Test Setup . . . . . . . . . . . . 36
1-21. Setup for Power Meter Calibration on Test Port 2 . . . . . . . . . .. . 37
1-22. Setup for Test Port 1 Input Frequency Response . . . . . . . . . . . .. 38
1-23. Setup for Power Meter Calibration on Test Port 1 . . . . . . . . . . . . . . 39
1-24. Test Port 2 Input Frequency Response Test Setup . . . . . . . . . . . . . . 40
1-25. Test Port Crosstalk Test Setup . . . . . . . . . . . . . . aaa a aa 41
1-26. HP 8753E Bottom View .. .. 1.2 2 41 41 LL 11 1111110 44
1-27. First Full 2-Port Calibration Test Setup . . .. 1.212221 11011110 45
1-28. Transmission Calibration Test Setup . . . . . . . . . . . . . . . . ... 46
1-29. Second Full 2-Port Calibration Test Setup . . . . . . . . . . . . . ... 48
1-30. Transmission Calibration Test Setup . . . . . . . . . . . . . . . . . .. .. 49
1-31. System Trace Noise Test Setup . . . . . . . . . . . . . . e ee... 50
1-32. System Trace Noise Test Setup . . . . . . . . . . . . . . . . . . ... 52
1-33. S11 1-Port Cal Test Setup . . . . . . . 1 111111 1010 11 111110 56
1-34. Test Port 2 Input Impedance Test Setup . . . . . . . . . . . . . . ... 57
1-35. S22 1-Port Cal Test Setup . . . . . . . . aa ea ae e aaa a 57
1-36. Test Port 1 Input Impedance Test Setup . . . . . . . . . . . o... a aa | 58
1-37. Power Meter Calibration for Magnitude Dynamic Accuracy . . . . . . . . 62
1-38. Full 2-Port Calibration with Adapter Removal . . . . . . . . . . . ... 64
1-39. Magnitude Dynamic Accuracy Measurement . . . . . . . . . . . . . .. 66
1-40. Test Port Magnitude Compression Test Setup . . . . . . . . . . . .. .. . 69
1-41. Test Port Phase Compression Test Setup . . . . . . . . . . . . . . .... 72
1-42. Test Port Output Harmonics Test Setup . . . . . . . . . . . . . .... 75
1-43. Receiver Harmonics Test Setup . . . . . . . . . . . . « . . . 76
1-44. Test Port Output Harmonics Test Setup . . . . . . . . . . . . . .. .. 79
1-45. Receiver Harmonics Test Setup . . . . . . . . . . . . . . . . .. ... 81
Tables
1-1. Magnitude Dynamic Accuracy Calculations
т & 2 & & » ко a a a ow
61
System Verification and
Performance Tests
There are two ways to confirm that the HP 8753E network analyzer is able to make
measurements as specified by Hewlett-Packard. Both ways are described in this chapter.
The first way is through the system verification procedure. With system verification, the
performance of the network analyzer is confirmed as a complete measurement system. The
network analyzer is used to measure the traceable behavior of test devices that are part of
a verification kit. All the measurement uncertainties of the network analyzer, taken as a
measurement system, have been accounted for in the serialized data disk shipped with the
verification Kit.
The specified performance of the network analyzer can also be confirmed using the series
of performance tests described in this chapter. Successful completion of an individual test
confirms the specified performance of the specific subsystem tested, such as the source or
receiver. Successful completion of the whole series of performance tests confirms the specified
performance of the network analyzer as a complete measurement system.
System Verification and
Performance Tests
1
How to Test the Performance of Your Analyzer o
To obtain the same quality of performance testing that Hewlett-Packard has administered at the —
factory, you must perform: |
u the system verification procedure
OR
u all of the performance test procedures.
This quality of performance testing guarantees that the analyzer is performing within all of
the published specifications. Hewlett-Packard will issue a Certificate of Calibration for your
analyzer if two conditions are met:
1. Your analyzer passes all the performed tests. —
2. The equipment and standards that you used to perform the tests are traceable to a national
standards institute.
Note If you have a particular type of measurement application that does not
use all of the analyzer's measurement capabilities, you may ask your local
Hewlett-Packard Customer Service Center for a subset of specifications that
you Want verified. However, this does create a potential for making incorrect
measurements, by using a different application than what was specified.
Sections in this Chapter
= System Verification
u Performance Tests
1. Test Port Output Frequency Range and Accuracy
2. External Source Mode Frequency Range
3. Test Port Output Power Accuracy
4. Test Port Output Power Range and Linearity
5. Minimum R Channel Level
6. Test Port Input Noise Floor Level
7. Test Port Input Frequency Response
8. Test Port Crosstalk
9. Calibration Coefficients
10. System Trace Noise (Only for Analyzers without Option 006)
11. System Trace Noise (Only for Analyzers with Option 006)
12. Test Port Input Impedance
13. Test Port Receiver Magnitude Dynamic Accuracy
14. Test Port Receiver Magnitude Compression
15. Test Port Receiver Phase Compression
16. Test Port Output/Input Harmonics (Option 002 Analyzers without Option 006 only)
17. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 only)
2 System Verification and
Performance Tests
Performance Test Record
Find and use the appropriate “Performance Test Record” in the following subchapters:
m Performance Test Record for 30 kHz to 3 GHz
a Performance Test Record for 30 kHz to 6 GHz
System Verification and 3
Performance Tests
System Verification Cycle and Kit Re-certification
Hewlett-Packard recommends that you verify your network analyzer measurement system
every six months. Hewlett-Packard also suggests that you get your verification kit re-certified
annually. Refer to HP 85029B 7-mm Verification Kit Operating and Service Manual for more
information.
Note The system verification procedures can also apply to analyzers with Option 075
(75 ohm analyzers) if minimum loss pads and type-N (m) to APC-7 adapters are
used.
Check to see how the verification kit floppy disk is labeled:
a If your verification disk is labeled HP 8753D Verification Data Disk, or
HP 8753D & HP 8753E Verification Data Disk, you may proceed with the system
verification.
m If your verification disk is not labeled as indicated above, you may send your HP 85029B
7-mm verification kit to the nearest service center for recertification, which includes a data
disk that you can use with the HP 8753E.
4 System Verification and
Performance Tests
HP 8753E System Verification
Equipment Required
Calibration Kit, 7-mm .................. PSS
Verification Kit, 7-mm ................... PS
Test Port Extension Cable Set, 7-mm ..........................
Printer ............. a noria La aa a
Additional Equipment Required for Option 075 Analyzers
Minimum Loss Pad (2), 50 Qto 75 Q ................... ee
Adapter (2), APC-7 to Type-N (mM) .............................
Analyzer warmup time: 1 hour
This system verification consists of three separate procedures:
1. Initialization
2. Measurement Calibration
3. Device Verification
Initialization
PS …... HP 85031B
TOS HP 85029B
O HP 11857D
HP ThinkJet/DeskJet/LaserJet
ens ceeieeoo....... HP 11852B
ES HP 11525A
1. Connect the equipment as shown in Figure 1-1. Let the analyzer warm up for one hour.
HP 8753E
NETWORK ANALYZER
HP-IB OR
SERIAL OR
PARALLEL
PRINTER
НР 118575
CABLE SET
CABLE 1 CABLE 2
REFERENCE TEST | ana. REFERENCE TEST
PORT 1 PORT 2
sgble
Figure 1-1. System Verification Test Setup
2. While the equipment is warming up, review the “Connector Care Quick Reference”
information in the “Service Equipment and Analyzer Options” chapter. Good connections
and clean, undamaged connectors are critical for accurate measurement results.
3. Insert the verification kit disk into the analyzer disk
4. Press (Preset ) (Save/Recall ) |
5. If
ou want a printout of the erific ti
drive.
data for all the devices, press
System Verification and 5
Performance Tests
Note If you switch on the record function, you CANNOT switch it off during the
verification procedure.
6. Position the paper in the printer so that printing starts at the top of the page.
7. If you have difficulty with the printer:
m If the interface on your printer is HP-IB, verify that the printer address is set to 1 (or
change the setting in the analyzer to match the printer).
m If the interface on your printer is serial or parallel, be sure that you selected the printer
port and the printer type correctly (refer to the HP 8753E Network Analyzer User's
Guide for more information on how to perform these tasks).
8. Press |
9. The analyzer displays Sys Ver Init DONE, the initialization procedure is complete.
Caution DO NOT press or recall another instrument state. You must use the
instrument state that you loaded during the initialization procedure.
Measurement Calibration
10. ;
Press
11.
12.
13. Connect the “open” end of the open/short combination (supplied in the calibration kit) to
reference test port 1, as shown in Figure 1-2.
HP 8753E
NETWORK ANALYZER
a — ho
(89591 HP-IB OR
OO
DO
a
0000 PARALLEL PRINTER
Ë E —
e | )
НР 118570
CABLE SET
0 [=== |
0000 0000
088 g
on? 088 6
o
D
o 8
CABLE 1 CABLE 2
REFERENCE TEST REFERENCE TEST
PORT 1 PORT 2
Mya dem N
i | 0 i i E
OPEN SHORT LOAD OPEN SHORT LOAD
sybZe
Figure 1-2. Connections for Measurement Calibration Standards
6 System Verification and
Performance Tests
ra
14.
15.
16.
17.
18.
19.
21.
22.
23.
24.
25.
20.
27.
28.
29.
30.
Press
When the analyzer finishes measuring the standard, connect the “short” end of the
open/short combination to reference test port 1.
Press
When the analyzer finishes measuring the standard, connect the 50 ohm termination
ibration kit) to reference test port 1.
When the analyzer finishes measuring the standard, connect the “open” end of the
open/short combination to reference test port 2.
When the analyzer finishes measuring the standard, connect the “short” end of the
open/short combination to reference test port 2.
Press
When the analyzer finishes measuring the standard, connect the 50 ohm termination to
reference test port 2.
Press
When the analyzer finishes measuring the standard, press ST:
The analyzer briefly displays COMPUTING CAL COEFFICIENTS.
Connect the test port cables as shown Figure 1-3.
HP 8753E
NETWORK ANALYZER
PORT 2
HP 118570
EXTENSION CABLE SET
Figure 1-3. Transmission Calibration Setup
Press (Save/Recall)
calibration into the analyzer internal m
y.
When the analyzer finishes saving the instrument state, press
sgble
to save the
System Verification and 7
Performance Tests
Device Verification = a
31. press & ee DO ; —
32. At the prompt, connect the 20 dB attenuator (supplied in the verification kit) as shown in НЕ
Figure 1-4.
33. Press CONTI NU > to run the test: —
® If you switched OFF the record function, you have to press C E after each
S-parameter measurement. —
m If you switched ON the record function, the analyzer measures all S-parameters
(magnitude and phase) without pausing. Also, the analyzer only displays and prints the —
PASS/FAIL information for the S-parameter measurements that are valid for system
verification.
HP 8753E
NETWORK ANALYZER
al co dana
a o
o
o mins
E
Ca —
VERIF ICATION DEVICE —
sgbde
Figure 1-4. Connections for the 20 dB Verification Device
34. When the analyzer finishes all the measurements, connect the 50 dB attenuator (supplied in
the verification kit), as shown in Figure 1-5.
HP 8753E
NETWORK ANALYZER
oo DOQQ
9000
oo 09000
509 0000
o бо о
© 655
0000 0090
00
wn] PORT 2
M50 de ATTEN
[О
VERIFICATION DEVICE
sgbSe
Figure 1-5. Connections for the 50 dB Verification Device
35. Press (f) -
8 System Verification and
Performance Tests
36. When all measurements are complete, replace the verification device with the verification
mismatch, as shown in Figure 1-6. Be sure that you connect Port A of the verification
`` mismatch to reference test port 1.
HP 8753E
NETWORK ANALYZER
co 00060
0000 ooob
PORT A PORT B
VERIFICATION MISMATCH
sgbbe
Figure 1-6. Mismatch Device Verification Setup
38. When the analyzer finishes all the measurements, connect the mismatch verification device,
as shown in Figure 1-7. Notice that Port B is now connected to reference test port 1.
HP 8753E
NETWORK ANALYZER
aa 5000
o
0000 0000
PORT В
PORT A
VERIFICATION MISMATCH
sgb/e
— Figure 1-7. Mismatch Device Verification Setup
39. Press |
40. You have completed the system verification procedure when the analyzer displays Ver Def
4 DONE.
System Verification and 9
Performance Tests
In Case of Difficulty
1. Inspect all connections. DO NOT disconnect the cables from the analyzer test ports. Doing
so WILL INVALIDATE the calibration that you have done earlier.
2. Press Save/Recall) $
3. Repeat the “Device Verification” procedure.
4. If the analyzer still fails the test, check the measurement calibration as follows:
a. Press (Preset).
- Recall the calibration by pressing (Save/Recall : .
¢. Use the front panel knob to highlight the calibration you want to recall and press
e
d. Connect the short to reference test port 1.
g. Check that the trace response is 0.00 + 0.05 dB.
h. Disconnect the short and connect it to reference test port 2.
i. Press (Meas) В
J. Check that the trace response is 0.00 + 0.05 dB.
k. If any of the trace responses are out of the specified limits, repeat the “Measurement
Calibration” and “Device Verification” procedures.
5. Refer to the “Start Troubleshooting Here” chapter of the HP 8753E Service Guide for more
troubleshooting information. |
10 System Verification and
Performance Tests
1. Test Port Output Frequency Range and Accuracy
Specifications
Frequency Range | Frequency Accuracy!
—30 kHz to 3 GHz +10 ppm
3 GHz to 6 GHZ? +10 ppm
1 At 25° C +5° C.
2 Only for analyzers with Option 006 — 30 kHz to
6 GHz range.
Required Equipment
Frequency Counter (30 kHz to 500 Mhz) .................................... HP 5350B/51B/52B
Frequency Counter (500 MHzto 6 GHz) ......................... e... ea HP 5350B/51B/52B
Cable, 50Q Type-N, 24-inch ........ aa aanaaa eco. HP P/N 8120-4781
Adapter, APC-3.5 (f) to Type-N (f) aa aaaaeco. HP P/N 1250-1745
Adapter, APC-7Tto Type-N(f) .......... iii... PS HP P/N 11524A
Adapter, Type-N (f) to BNC (Mm) ........... i, HP P/N 1250-1477
Additional equipment needed for an HP 8753E with Option 075
Minimum Loss Pad, 502to 75Q ............................. RD . HP 11852B
Analyzer warmup time: 30 minutes
Perform this test to verify the frequency accuracy of the HP 8753E over its entire operating
frequency range.
System Verification and 11
Performance Tests
1. Connect the equipment as shown ir. Figure 1-8.
HP 8753E HP 5343A
NETWORK ANALYZER MICROWAVE
WITH OPTION 075 FREQUENCY COUNTER
i?” Bn,
В Sees BEES Ге |
= весе, 385 ва =
9 992 Go о co 2 0650 q > 7
Ф OQ PORT2 ** + x
= nd ADAPTER ADAPTER
PORT1 | * TYPE—N(f) APC—3.5 (f)
TO BNC (m) TO TYPE—N(f)
50% / 759
MIMIMUM
LOSS PAD
500 TYPE-N CABLE ASSEMBLY
| НР 534 ЗА
HP 8753E M | CROWAVE
NE TWORK ANALYZER FREQUENCY COUNTER
e]
=D | |
5 QH9 900080 900 000 ‚м:
ae pago 335 858 = -
| 8| 85 898 © SEXE a
= — i
P (® PORT2 x | x 1
= = ADAPTER ADAPTER
PORT! | * TYPE-N(1) APC-3.5 (f)
| TO BNC (m) TO TYPE—N(f)
ADAPTER
APC—7 TO
TYPE — N(f)
50 TYPE-N CABLE ASSEMBLY
* DIRECT CONNECTION sgbBe
Figure 1-8. Test Port Output Frequency Range and Accuracy Test Setup
2. Press с
3. Press (30) (k/m) and write the frequency counter reading on the “Performance Test Record.”
4. Repeat step 3 for each instrument frequency listed in the “Performance Test Record.”
In Case of Difficulty
1. If any measured frequency is close to the specification limits, check the time base accuracy
of the counter used.
2. If the analyzer fails by a significant margin at all frequencies (especially if the deviation
increases with frequency), the master time base probably needs adjustment. In this case,
refer to the “Frequency Accuracy Adjustment” procedure, located in the “Adjustments and
Correction Constants” chapter of the HP 8753E Service Guide. The “Fractional-N Frequency
Range Adjustment” also affects frequency accuracy.
3. Refer to the “Source Troubleshooting” chapter of the HP 8753E Service Guide for related
troubleshooting information.
12 System Verification and
Performance Tests
2. External Source Mode Frequency Range
Specifications
Frequency Range
30 kHz to 3 GHz
30 kHz to 6 GHz!
1 Only for analyzers with
Option 006 — 30 kHz to
6 GHz range.
Equipment Required
External Source ................. a HP 83620A
Cable, APC-7, 24-inch... aaa oereaaca. .... HP P/N 8120-4779
Adapter, APC-3.5 (f) to APC-7 .................... Ce ee LL... HP P/N 1250-1747
Adapter, APC-3.5 (m) to APC-7 .... e ea HP P/N 1250-1746
Analyzer warmup time: 30 minutes
Perform this test to verify that the analyzer’s reference channel, input R, is capable of phase
locking to an external CW signal.
1. On the external source, press (MHz/ psec) (POWER LEVEL) (=/<) GHz/dB(m)).
2. Connect the equipment as shown in Figure 1-9.
HP B753E
NETWORK ANALYZER SANTOR TOR
“a” ,
glee (gees e =
5 GAS о 0900 ss Fou
5/88 558 ¢ === oa sage
= . coco 8200000 oooo oca
© ® ana 0000 =
=a
R CHANNEL IN | % RF OUTPUT
ADAPTER ADAPTER
APC-3.5 (m) APC—3.5 (Е)
TO APC—7 TO APC-7
CABLE APC—7 24 INCH
* DIRECT CONNECTION sg69e
Figure 1-9. External Source Mode Frequency Range Test Setup
3. On the network analyzer, press |
4. Press (© M7).
5. Check to see if the analyzer is phase locking to the external CW signal:
m If the analyzer displays any phase lock error messages, write “unlock” in the
“Performance Test Record” for the set CW signal.
System Verification and 13
Performance Tests
m If the analyzer does not display any phase lock error messages, write “lock” in the
“Performance Test Record” for the set CW signal.
6. On the external source, press (MHz/ ut).
7. On the analyzer, press (20) (M/u).
8. Repeat step 5 through 7 for the other external source CW frequencies listed in the
“Performance Test Record.”
In Case of Difficulty
1. Be sure the external source power is set within 0 to —25 dBm.
2. Make sure the analyzer's “Ext Source Auto” feature is selected. In addition, verify that the
analyzer is set to measure its input channel R.
3. Verify that all connections are tight.
14 System Verification and
Performance Tests
3. Test Port Output Power Accuracy
Specifications
Frequency Range
Test Port Output
Power Accuracy!
300 kHz to 3 GHz
+1.0 dB
3 GHz to 6 GHz?
+1.0 dB
1 At 0 dBm and 25° C +5° C
2 Only for analyzers with Option 006 — 30 kHz
to 6 GHz range.
Equipment Required for 501 Analyzers
Po W er Meter ou; "'‘.. #; * *# à #4 Hé à #4 9 2 VV EEE EU E EE FF AE E EE EEE EE EE ETT 8
Power Sensor + = # k= > + 4% + tb 4 vu + + x BE # + ow X uv u =u 7 = # # #4 #4 = + #6 œ EE € ù = = + лок # 8 € oF % # #6 # 6 # + + 6 = = vu Ow + = E vu 4 # 4% 0 E = +4 VU 5 =
Additional Equipment Required for Analyzers with Option 006
Power Sensor ....................... Ce Cen ee oca aora, ss.
Equipment Required for 752 Analyzers
Power Meter = «#8 dé 8 PEE PAE BEE 1 4 CU 8 6 = 4 4 En # à # # WO + # # F + + + F * ¥ # uw + 2 = > *
Power Sensor 4 4 a © #4 # # # 6 5 + 4 4 еж 4 % 4 4 = a BA MOE OE ESA AE EA жк * * 4% 4 OEA
Analyzer warmup time: 30 minutes
Perform this test to confirm the accuracy of the HP 8753E source output power.
System Verification and
Performance Tests
HP 436A/437B/438A
HP 8482A
HP 11524A
HP 8481A
HP 436A/437B/438A
HP 8483A Option H03
15
1. Zero and calibrate the power meter. For more information of how to perform this task, refer
to the power meter operating manual.
er
2. Connect the equipment as shown in Figure 1-10.
HP 8753E
NETWORK ANALYZER ee
HP 438A —
Al IB
ADAPTER APC—7 to N(F)
—_
| nn
HP B481A HP B482A
POWER SENSOR POWER SENSOR
(OPTION 006 ONLY)
J
J
HP 8753E
NETWORK ANALYZER
WITH OPTION 075 I
pu” Net. —
8122 (8338
D000
5/26 830000 HP 438A
9 do 888 © AIN TT
pom _
— + — —
PORT 1 PORT 2
HP 8483A
POWER SENSOR
(OPTION H03)
s9610e
Figure 1-10. Source Output Power Accuracy Test Setup
8. Press (Preset). QU À OT
Note The factory preset test port power is 0 dBm.
4. Press | l (k/m). Set the calibration factor on the power meter for thisCW ~~ —
frequency.
5. Write the power meter reading on the “Performance Test Record.”
6. Repeat steps 4 and 5 for each CW frequency listed in the “Performance Test Record.” For
analyzers with Option 006, use the HP 8481A power sensor for all frequencies above 3 GHz.
16 System Verification and
Performance Tests
In Case of Difficulty
1. Be sure the source power is switched on. Press
power.
2. Refer to the “Source Troubleshooting” chapter of the HP 8753E Service Guide for more
troubleshooting information.
System Verification and
Performance Tests
17
4. Test Port Output Power Range and Linearity
Specifications
Power Range |Power Level Linearity!
—15 to +5 dBm +0.2 dB
+5 to +10 dBm +0.5 dB
1 Relative to 0 dBm output level.
Required Equipment
Power Meter ......... ee cave adócececo HP 437B/438A
Power Sensor ........................... ainda eder, LL La a La La LAVE LL ass ac HP 8482A
Adapter, APC-7 to Type-N (f) .............. .................. adas O HP 11524A
Additional Required Equipment for Analyzers with Option 006
Power Sensor 2. e eecidroceaoo LL Le a a a Ada A aa aa HP 8481A
Additional Required Equipment for Analyzers with Option 075
Power Sensor 2202000444 aa aa aa aa ed a ea ad a ae a a aa HP 8483A Option H03
Analyzer warmup time: 1 hour
Perform this test to verify the analyzer’s test port output power range and power level
linearity at selected CW frequencies.
18 System Verification and
Performance Tests
. Zero and calibrate the power metcr. Refer to the power meter operating and service
manual for more information on how to do this task.
. On the network analyzer, press С
calibration factor for this CW frequency.
; (k/m). Set the power meter
. Connect the equipment as shown in Figure 1-11.
HP 8753E
NETWORK ANALYZER
HP 438A
ADAPTER APC-7 to N(F)
LA se
í \
HP B481A HP 8482A
POWER SENSOR POWER SENSOR
3 — 6GHz 30KHz — 3GHz
/
J
HP 8753E
NETWORK ANALYZER
WITH OPTION 075
HP 438A
TN
PORT 1| * PORT 2
HP 8483A
POWER SENSOR
(OPTION HO)
* DIRECT CONNECTION
sgblle
Figure 1-11. Test Port Output Power Range and Accuracy Test Setup
. On the HP 438A, press (REL). This sets the current power level for relative power
measurement.
. On the network analyzer, press POWER
. Write the power meter reading in the “Results Measured” column on the “Performance Test
Record.”
. Calculate the difference between the analyzer test port power (which appears on the
analyzer’s display) and the power meter reading. Write the result in the “Power Level
Linearity” column on the “Performance Test Record.”
System Verification and
Performance Tests
10.
11.
12.
13.
14.
. Repeat steps 5 through 7 for the other power levels listed in the “Performance Test
Record.”
Set the power meter calibration factor for this CW frequency and press to set the
reference at this new frequency.
Press I
Write the power meter reading in the “Results Measured” column on the “Performance Test
Record.”
Calculate the difference between the analyzer test port power and the power meter
reading. Write the result in the “Power Level Linearity” column of the “Performance Test
Record.”
Repeat steps 11 through 13 for the other power levels listed in the “Performance Test
Record.”
In Case of Difficulty
1.
20
Ensure that the power meter and power sensor(s) are operating to specifications. Be sure
you set the power meter calibration factor for the CW frequency that you are testing.
Repeat this performance test.
System Verification and
Performance Tests
5. Minimum R Channel Level
Specifications
Frequency Range | Minimum R Channel Level
300 kHz to 3 GHz < —35 dBm
3 GHz to 6 GHz! <—30 dBm
1 Only for analyzers with Option 006 — 30 kHz to 6 GHz
range.
Required Equipment for 500 Analyzers
Adapter, APC-3.5 (m) to APC-7 aaa aaa HP P/N 1250-1746
Cable, APC-7 24-inch .......... 2 aaa ea ВИ HP P/N 8120-4779
Required Equipment for 75Q Analyzers (Option 075)
Minimum Loss Pad, 509 to 759 ........... UPR HP 11852B
Cable, 500 Type-N, 24-inch ............ eee LL aa aa LL La a HP P/N 8120-4781
Adapter. APC-3.5 (m) to Type-N (f) PS HP P/N 1250-1750
Analyzer warmup time: 1 hour
Perform this test to determine the minimum R channel input power level at which phase lock
can be accomplished.
1. Connect the equipment as shown in Figure 1-12.
HP 8753E
HP 8753E NETWORK ANALYZER
NETWORK ANALYZER WITH OPTION 075
pl ma. pd a
8122 Og
81765 585280
© Jo 288 ©
= o Da
® © a5 R CHANNEL
PORT1 PORT2 x PORT t PORT2 *
ADAPTER 500/750
[] APC-3.5 (m) MINIMUM ADAPTER
TO APC—7 LOSS PAD
CABLE APC—7 24 INCH
syb1Ze
* DIRECT CONNECTION .
Figure 1-12. Minimum R Channel Level Test Setup
2. Press (Ereset) (Mess) |
3. Press | и) POR
4. Press ]
System Verificationand 21
Performance Tests
The analyzer displays the message CAUTION: NO IF FOUND: CHECK R INPUT LEVEL.
7. Press (1) to increase the test port power by 1 dBm.
8. If the analyzer displays a phase lock error message, continue increasing the test port power
until phase lock is achieved.
9. Write the test port power, that is displayed on the analyzer, on the “Performance Test
Record.”
10. Repeat steps 5 through 9 for the other CW frequencies listed in the “Performance Test
Record.”
22 System Verification and
Performance Tests
— In Case of Difficulty
ma” 1. Check the flexible RF cable (W8, as shown in Figure 1-13) between the R sampler assembly
(A4) and the A11 phase lock assembly. Make sure it is connected between A11J1 (PL IF IN)
and 1st IF Out.
Caution Do not push cable W8 down next to the A11 phase lock assembly.
А10 АВ AN A12
W13
AU
wa
soôBée
Figure 1-13. Flexible RF Cable Location
System Verification and 23
Performance Tests
2. Using an ohmmeter, verify that the RF cable is not open. In addition, examine both the
cable connectors — measure the resistance between the cable center pin and the cable
connector and make sure it is not close to zero.
3. Check the R sampler by substituting it with the B sampler (A6).
a. Move cable W8 to the B sampler (A6), as shown in Figure 1-14.
AN
| A6_
LAS |
O
Qeee в | Г”
sgó11Se
Figure 1-14. Connections for Substituting the R Sampler (A4)
4. Connect the equipment as shown in Figure 1-15.
HP 8753E
NETWORK ANALYZER
{ § ==] \
00009 0000
0
D
D
D
о
„и 2U—INCH CABLE
sg6l3e
Figure 1-15. Setup for Checking the R Sampler (A4)
Use the following specifications:
300 kHz to 3 GHz <—27 dBm
J GHZ to 6 GHz <-22 dBm
24 System Verification and
Performance Tests
6. If the analyzer fails the test, replace the Al1 assembly.
7. Verify that the high/low band adjustments are still within specifications. For more
information on how to perform this task, refer to the “High/Low Band Transition
Adjustment” located in the “Adjustments and Correction Constants” chapter of the
HP 8753E Service Guide.
8. Refer to the “Source Troubleshooting” chapter of the HP 8753E Service Guide for more
troubleshooting information.
System Verification and 25
Performance Tests
6. Test Port Input Noise Floor Level
Specifications
Frequency Range | Test Port | ТЕ Bandwidth | Average
Noise Level
300 kHz to 3.0 GHz | Port 1 3 kHz —82 dBm
300 kHz to 3.0 GHz | Port 1 10 Hz —102 dBm
300 kHz to 3.0 GHz | Port 2 3 kHz —82 dBm
300 kHz to 3.0 GHz | Port 2 10 HZ —102 dBm
3.0 GHz to 6.0 GHZ! | Port 1 3 kHz —77 dBm
3.0 GHz to 6.0 GHz! | Port 1 10 Hz —97 dBm
3.0 GHz to 6.0 GHZ! | Port 2 3 kHz —77 dBm
3.0 GHz to 6.0 GHZ! | Port 2 10 Hz —97 dBm
1 Only for analyzer with Option 006 - 30 kHz to 6 GHZ range.
Equipment Required for 500 Analyzers
Calibration Kit, 7-mm .............. 2. ana aa ae a ea a a ee aan RE
Equipment Required for 750 Analyzers
Calibration Kit, Type-N ......... PAL AAA AA AAA RAA A Rae ee ee aa a aa a aa a an
Analyzer warmup time: 1 hour
HP 850318
HP 85036B
Perform this test to determine the HP 8753E port 1 and port 2 noise floor levels at the input
test ports.
25 System Verification and
Performance Tests
— Port 1 Noise Floor Level from 300 kHz to 3 GHz (IF BW = 3 kHz)
= 1. Connect the equipment as shown in Figure 1-16.
HP 8753E
NETWORK ANALYZER
500 TERMINATIONS
sq614e
Figure 1-16. Source Input Noise Floor Test Setup
— 2. Press (Preset) (Avg) IF BW (3000) (2) (Menu) #
3. Press (Veas) 1 (Format) (Sele Ref) 4 -
5. When the analyzer finishes the sweep, notice the mean value (which appears on the
analyzer display).
6. Convert the measured linear magnitude mean value to log magnitude, using this equation.
Power (dBm) = 20 * [log, (linear magnitude mean value)]
a Note Notice that the mean value that is displayed on the analyzer is in jUnits. So,
for example, if the displayed value is 62 “U, the value that you would put in
the equation is (62 x 106).
7. Write this calculated value on the “Performance Test Record.”
System Verification and 27
Performance Tests
Port 1 Noise Floor Level from 300 kHz to 3 GHz (IF BW = 10 Hz)
8. Press
7. Press TRI
10. When the analyzer finishes the sweep, notice the mean value.
: to change the IF bandwidth to 10 Hz.
11. Convert the measured linear magnitude mean value to log magnitude, using this equation.
Power (dBm) = 20 x [log, (linear magnitude mean value)]
12. Write this calculated value on the “Performance Test Record.”
Port 2 Noise Floor Level from 300 kHz to 3 GHz (IF BW = 10 Hz)
14. Press (fam) *
15. When the analyzer finishes the sweep, notice the mean value.
16. Convert the measured linear magnitude mean value to log magnitude, using this equation.
Power (dBm) = 20 [log; (linear magnitude mean value)]
17. Write this calculated value on the “Performance Test Record.”
Port 2 Noise Floor Level from 300 kHz to 3 GHz (IF BW = 3 kHz)
18. Press (Ave)
19. Press 1
20. When the analyzer finishes the sweep, notice the mean value.
O) 77) to change the IF bandwidth to 3 kHz.
21. Convert the measured linear magnitude mean value to log magnitude, using this equation.
Power (dBm) = 20 * [log, ,(linear magnitude mean value)]
22. Write this calculated value on the “Performance Test Record.”
23. This completes the “Test Port Input Noise Floor Level” procedure if your analyzer does not
have Option 006. Otherwise continue with the next section.
Port 2 Noise Floor Level from 3 GHz to 6 GHz (IF BW = 3 kHz)
24. Press (Start) (3) (G/n) (Stop) (6) (G/r).
25. Press
26. When the analyzer finishes the sweep, notice the mean value.
27. Convert the measured linear magnitude mean value to log magnitude, using this equation.
Power (dBm) = 20 « [log, (linear magnitude mean value
28. Write this calculated value on the “Performance Test Record.”
28 System Verification and
Performance Tests
Port 2 Noise Floor Level from 3 GHz to 6 GHz (IF BW = 10 Hz)
29, Ва
30.
31. When the analyzer finishes the sweep, notice the mean value.
32. Convert the measured linear magnitude mean value to log magnitude, using this equation.
Power (dBm) = 20 « [log (linear magnitude mean value)]
33. Write this calculated value on the “Performance Test Record.”
Port 1 Noise Floor Level for 3 GHz to 6 GHz (IF BW = 10 Hz)
35. Press (Menu)
36. When the analyzer finishes the sweep, notice the mean value.
37. Convert the measured linear magnitude mean value to log magnitude, using this equation.
Power (dBm) = 20 * [log, (linear magnitude mean value)]
38. Write this calculated value on the “Performance Test Record.”
Port 1 Noise Floor Level from 3 GHz to 6 GHz (IF BW = 3 kHz)
39. Press W
40. Press (Menu)
41. When the analyzer finishes the sweep, notice the mean value.
42. Convert the measured linear magnitude mean value to log magnitude, using this equation.
Power (dBm) = 20 [log (linear magnitude mean value)]
43. Write this calculated value on the “Performance Test Record.”
In Case of Difficulty
1. Perform the “ADC Linearity Correction Constants (Test 52),” located in the “Adjustments
and Correction Constants” chapter of the HP 8753F Service Guide.
2. Repeat the “Test Port Input Noise Floor Level” procedure.
3. Suspect the A10 Digital IF assembly if the analyzer fails both test port input noise floor tests.
4. Refer to the “Receiver Troubleshooting” chapter of the HP 8753E Service Guide for more
troubleshooting information.
System Verification and 29
Performance Tests
7. Test Port Input Frequency Response
Specifications
Frequency Range Test Port Input
Frequency Response
300 kHz to 3 GHz | Port 1 | +1 dB
300 kHz to 3 GHz | Port 2 +1 dB
3 GHz to 6 GHZ! | Port 1 +2 dB
3 GHz to 6 GHz! | Port 2 +2 dB
1 Only for analyzers with Option 006 — 30 kHz to 6 GHz range.
Equipment Required for 500 Analyzers
Power Meter o.oo aa aa aa aa aan a nana aa ana HP 436A/437B/438A
Power Sensor .................. iin. ee HP 8482A
Cable, APC-T7 24-inch ae aoaaaaaoo oa HP P/N 8120-4779
Adapter, APC-7 to Type-N (£) ........ aaa SP HP 11524A
Additional Equipment Required for Analyzers with Option 006
Power Sensor ...................... ee ee A AA da de ee AA ae a a an aa aa a HP 8481A
Equipment Required for 750 Analyzers
Power Meter ......... aerea enero a eran recrea oeaoocca. HP 436A/437B/438A
Power Sensor ........ SS ива ооов он ововнное HP 8483A Option HO3
Cable, Type-N .............. Cee KR HP P/N 8120-2408
Analyzer warmup time: 1 hour
Perform this test to examine the vector sum of all test setup error vectors in both magnitude
and phase change as a function of frequency.
30 System Verification and
Performance Tests
- Press
Power Meter Calibration for Test Port 1 from 300 kHz to 3 GHz
1. Zero and calibrate the power meter.
2. Connect the equipment as shown in Figure 1-17.
HP 8753E
NETWORK ANALYZER
Te
DO 0000
888 590000 HP 43
438A
со 588 © » |
ao o HP-IB — ra
O = «=>
ЧН
FORT 2
CABLE
АРС-7 ADAPTER HP 8482A
POWER SENSOR
HP B753E
NETWORK ANALYZER
WITH OPTION 075
0000 HP 438A
750 CABLE HP B4B3A
TYPE N POWER SENSOR
(OPTION HO3)
` o!
Figure 1-17. Setup for Power Meter Calibration on Test Port 1
596156
. Press (Preset } (Start) (300) (k/m).
4. Only for Analyzers with Option 006: Press (G/n).
9. Press (Local) £
| t until the analyzer shows the correct power meter
model.
System Verificationand 31
Performance Tests
B. The default power meter HP-IB address is 13. Make sure it
is the s same as your power meter HP-IB address. Otherwise, use the analyzer front panel
keypad to enter the correct HP-IB address for your power meter.
8.
9. Press I on | : to turn the auto power range off.
Note The analyzer displays the PRm annotation, indicating that the analyzer power
range is set to MANUAL.
10. Press p 0! B . to uncouple the test port output power.
Refer to the back of the power sensor
to locate the different calibration factor values along with their corresponding frequencies.
Note The analyzer’s calibration factor sensor table can hold a MAXIMUM of 12
calibration factor data points.
The following softkeys are included in the sensor calibration factor entries menu:
press to select a point where you can use the front panel knob or
entry keys to enter a value.
press to edit or change a previously entered value.
press to delete a point from the sensor calibration factor table.
select this key to add a point into the sensor calibration factor table.
select this key to erase the entire sensor calibration factor table.
select this key when done entering points to the sensor calibration
factor table.
32 System Verification and
Performance Tests
13.
14.
15.
As an example, the following are the keystrokes for entering the first two calibration factor
data points for the HP 8482A power sensor (assuming CF% = 96.4 at 100 kHz and CF% =
98.4 at 300 kHz):
the front panel knob to point to the data point you want to modify and press El
Press the appropriate softkeys to create a power sensor calibration factors table.
ip to start the power meter calibration.
Wait until the analyzer finishes the sweep, then continue with this procedure.
Note The analyzer displays the PC annotation, indicating the power meter calibration
is done and the error correction is active.
System Verification and 33
Performance Tests
Test Port 2 Input Frequency Response from 300 kHz to 3 GHz
16. Connect the equipment as shown in Figure 1-18.
17.
18.
19.
20.
21.
34 System Verification and
Figure 1-18. Test Port 2 Input Frequency Response Test Setup
Press (Marker)
HP 8753E
NETWORK ANALYZER
``
co /\ 9000
TTT 24—INCH CABLE
maximum magnitude location of the trace.
sgbl3e
IN to put marker 1 at the
to position marker 2 at the
Write the marker 1 or marker 2 value (which appears on the analyzer display), whichever
has the larger absolute magnitude, in the “Performance Test Record.”
Performance Tests
Power Meter Calibration on Port 2 from 300 kHz to 3 GHz
22. Connect the equipment as shown Figure 1-19.
HP 8753E
NETWORK ANALYZER
HP 4538A
PORT 1 PORT 2
ADAPTER
АРС-7 to N(F)
CABLE APC-7
24 INCH
HP 8482A
POWER SENSOR
HP 8753E
NETWORK ANALYZER
WITH OPTION 075
ao06 HP 438A
0000 0000 |
do 588 9 TN
| в ео o HP-IB [lo
© —
ЧР — :
PORT 1 PORT 2 B
7501 CABLE HP 8483A
TYPE N POWER SENSOR
(OPTION HO3)
` OL |
sg6i6e
Figure 1-19. Setup for Power Meter Calibration on Test Port 2
23. Press 3
24. Press P JEEP to start the power meter calibration for
test port 2.
25. When the analyzer displays the message POWER METER CALIBRATION SWEEP DONE, connect
the equipment as shown as in Figure 1-20.
System Verification and 35
Performance Tests
HP 8753E
NETWORK ANALYZER
| 0000 0000
и 24—INCH CABLE
No
5961 5е
Figure 1-20. Test Port 1 Input Frequency Response Test Setup
Test Port 1 Input Frequency Response from 300 kHz to 3 GHz
26.
27.
28.
29.
30.
Press (Mess) 3
Press (Waker) (Varker Form) №
Press (Marker Fein) M SE
Write the marker 1 or marker 2 reading, whichever has the larger absolute magnitude, in
the “Performance Test Record.”
This completes the “Test Port Input Frequency Response” procedure if your analyzer does
not have Option 006. Otherwise continue with the next sections.
Power Meter Calibration for Test Port 2 from 3 GHz to 6 GHz
31.
32.
36
Replace the power sensor with the HP 8481A, and then setup the power meter:
Mm If the power meter is an HP 438A, press (LCL).
m If the power meter is an HP 437B, press (PRESET/LOCAL).
m If the power meter is an HP 436A, cycle the line power.
Connect the equipment as shown in Figure 1-21.
System Verification and
Performance Tests
HP 8753E
NETWORK ANALYZER
\
/
Зо | са сасо
151386 850000
191530 6950000 HP 438A
16/4 988 ° IN
15106 O HP IB ee
P - J |
PORT 1 PORT 2 В
ADAPTER
APC-7 to N(F)
CABLE APC—7
24 INCH
HP 8481A
POWER SENSOR
HP 8753E
NETWORK ANALYZER
WITH OPTION 075
HP 438A
0000 0000
HP B483A
POWER SENSOR
(OPTION H03)
` a]
Figure 1-21. Setup for Power Meter Calibration on Test Port 2
33. Press Sa) E) E) O ED).
34. Press
759 CABLE
TYPE N
sgbl/e
35. Press I . Repeat step 10 to build a calibration
factor sensor table for the HP 8481A power sensor.
36. Press DO E to exit the sensor calibration factor entries menu.
37.
38. Press RE EP to start the power meter calibration.
System Verification and 37
Performance Tests
Test Port 1 Input Frequency Response from 3 GHz to 6 GHz
39. When the analyzer finishes the calibration sweep, connect the equipment as shown in —
Figure 1-22.
HP 8753E
NETWORK ANALYZER
sg618e
Х to position marker 2 at the
maximum magnitude location of the trace.
43. Write the marker 1 or marker 2 reading, whichever has the largest absolute magnitude, in
the “Performance Test Record.”
38 System Verification and
Performance Tests
— Power Meter Calibration on Test Port 1 from 3 GHz to 6 GHz
— 44. Connect the equipment as shown in Figure 1-23.
HP 8753E
NETWORK ANALYZER
HP 43BA
„о
ADAPTER
APC—7 to N(F)
CABLE APC—7
24 INCH
HP 8481A
POWER SENSOR
HP 8753E
NETWORK ANALYZER
— WITH OPTION 075
pw J,
80/0 HP—IB E —]—]———
© © =|
_ EF Naa
PORT 1 PORT 2
750 CABLE HP B483A
— TYPE N POWER SENSOR
(OPTION HOJ)
` of]
ms 5956196
Figure 1-23. Setup for Power Meter Calibration on Test Port 1
P to start the power meter calibration for
System Verificationand 39
Performance Tests
Test Port 2 Input Frequency Response from 3 GHz to 6 GHz
47. When the analyzer displays the message POWER METER CALIBRATION SWEEP DONE, connect
the equipment as shown as in Figure 1-24.
HP B753E
NETWORK ANALYZER
TN
O =]
у
"uu
PORT 1
Figure 1-24. Test Port 2 Input Frequency Response Test Setup
48. Press (Meas) I
49. Press (Marker)
50. Press M
“Performance Test Record.”
In Case of Difficulty
1. Be sure you have used the correct power sensor for the frequency range.
2. Verify that the calibration factors that you have entered for the power sensors are correct.
(Marker Fctn) Marker Fctn) Mi
(Marker Fctn) EZ
51. Write the marker 1 or marker 2 reading, whichever has the largest magnitude, in the
0000 0000
P Ф
I
CABLE
24 INCH
3. Repeat this test with a “known good” thru cable.
40 System Verification and
Performance Tests
8. Test Port Crosstalk
Specifications
Frequency Range | Crosstalk!
300 kHz to 3 GHz 100 dB
3 GHz to 6 GHz? 90 dB
1 At 25° C +5° C.
2 Only for analyzers with Option 006 —
30 kHz to 6 GHz range.
Required Equipment for 500 Analyzers
Calibration Kit, 7-mm ............................ PS HP 85031B
Cable, APC-7 24-inch ........ aa a aa aa aa aa arereacera Le HP P/N 8120-4779
Required Equipment for 750 Analyzers
Calibration Kit, 750, Type-N
Analyzer warmup time: 1 hour
LS LE a RB a BE 4 4 4 4 4 5 4 4 4 4 = LL... HP 850308
Perform this test to verify the signal leakage between the analyzer’s test ports.
1. Connect the equipment as shown in Figure 1-25.
HP B753E
NETWORK ANALYZER
el”
ol oo 0000
DO сос A000
a 00 madDOO
= OO 650000
| = de Bas ©
5190 o
ро]
Ф Ф
ЧР чи
PORT 1 PORT 2
x | LT SHORT END
CABLE APC—7
OPEN/SHORT
4 /
24 INCH (500)
“e OPEN END
À DIRECT CONNECTION
HP 8753E
NETWORK ANALYZER
WITH OPTION 075
pi”
/
|
PORT 1
Ф
SHORT (т)
(750)
0000 0000
00°
00
00
00
o
*1 PORT 2
ADAPTER
TYPE-N(m o
PENA (50)
SHORT (f)
(750)
sg620e
Figure 1-25. Test Port Crosstalk Test Setup
2. Press
3. Pr ess | нала
System Verification and 41
Performance Tests
Crosstalk to Test Port 2 from 300 kHz to 3 GHz
4. Press (art) (0) CZ) Etes) ©) CA)
O. Press ( | I
9. Write the marker value (which appears on the analyzer display) in the “Performance Test
Record.”
Crosstalk to Test Port 1 from 300 kHz to 3 GHz
10. Press (Meas) 3
11. Press (Menu) °
12. Press (Marker Fan) M 3
13. Write the marker value (which appears on the analyzer display) in the “Performance Test
Record.”
14. This completes the “Test Port Crosstalk” performance test if your analyzer does not have
Option 006. Otherwise, proceed to the next section.
Crosstalk to Test Port 1 from 3 GHz to 6 GHz
15. Press (eae) (0) (on) (22) © (27)
17. Press M с.
18. Write the marker value (which appears on the analyzer display) in the “Performance Test
Record.”
42 System Verification and
Performances Tests
— Crosstalk to Test Port 2 from 3 GHz to 6 GHz
"a 1.
— 3. Press M
4. Write the marker value (which appears on the analyzer display) in the “Performance Test
Record.”
In Case of Difficulty
— 1. Remove the instrument top cover. Using an 8 Ib-inch torque wrench, verify that ALL
semirigid cables connected to the sampler/mixer assemblies are tight. In addition, tighten
any loose screws on the sampler/mixer assemblies (A4/5/6) and the pulse generator assembly
a (AT).
2. Remove the instrument bottom cover. Refer to Figure 1-26. Verify that cables W1, W31 and
__ W32 are tight.
3. Repeat this test.
System Verification and 43
Performance Tests
W32
44 System Verification and
Performance Tests
W1
Figure 1-26. HP 8753E Bottom View
W31
sg6102e
9. Calibration Coefficients
Specifications
Uncorrected! Frequency Range
Error Terms [300 kHz to 1.3 GHz| 1.3 GHz to 3 GHz| 3 GHz to 6 GHz?
Directivity 35 dB 30 dB 25 dB
Source Match 16 dB 16 dB 14 dB
Load Match 18 dB 16 dB 14 dB
Transmission Tracking +1.5 dB +1.5 dB +2.5 dB
Reflection Tracking +1.5 dB +1.5 dB +2.5 dB
1 At 25% +5% C, with less than 1° C deviation from the measurement calibration temperature.
2 Only for analyzers with Option 006 — 30 kHz to 6 GHz range.
Equipment Required for 509 Analyzers
Calibration Kit, 7-mm ................... Ce aa aaron ce ocean nadeaacón eo HP 85031B
Cable, APC-7, 24-inch ............ a aaa aaa HP P/N 8120-4779
Equipment Required for 750 Analyzers
Calibration Kit, Type-N ...... nena. ee ee Ce HP 850368
Cable, Type-N, 24-inch ............ aaa aoaaa aaa oaeca ... HP P/N 8120-4781
Analyzer warmup time: 30 minutes
Perform this procedure to verify the analyzer uncorrected test port characteristics.
Note The crosstalk calibration coefficients are omitted in this procedure. They are
covered in the “Test Port Crosstalk” performance test.
First Full 2-Port Calibration
1. Connect the equipment as shown in Figure 1-27.
HP 8753E
NETWORK ANALYZER
ogg ©
aa Doa
aa o
O come
ь
PORT 2
ANALYZER -
TEST PORT 1
CABLE 24 INCH
REFERENCE
TEST PORT 2 )
Figure 1-27. First Full 2-Port Calibration Test Setup
sgb2ie
System Verification and 45
Performance Tests
Го
- Press (кт). —
- Press (Cal) CAL KIT SELE |
. Replace the open/short combination with the 50 ohm termination (supplied in the
calibration kit). —
10. Press }
17. When the analyzer displays PRESS DONE’ IF FINISHED WITH STD(s), press
Wait for the message COMPUTING CAL COEFFICIENTS to disappear from the analyzer display
before proceeding to the next step.
18. Connect the equipment as shown in Figure 1-28.
HP 8753E
NETWORK ANALYZER
para a —
D ===]
0000 0000
D
D
DO
OD
DO
oO
CABLE
24 INCH
sg6l8Be TT
Figure 1-28. Transmission Calibration Test Setup
19. Press т o
46 System Verification and
Performance Tests
22. When the analyzer finishes the test, press (Marker).
23. Using the front panel knob, locate the maximum value of the data trace for the 300 kHz to
1.3 GHz frequency range.
24. Write the maximum value in the “Performance Test Record.”
25. Repeat the previous two steps for the other frequency range(s) listed on the “Performance
Test Record.”
Source Match (Forward) Calibration Coeflicient
26. Press (System) SERV TS
27. When the analyzer finishes the test, repeat steps 22 through 25.
Transmission Tracking (Forward) Calibration Coefficient
28. Press (System) $
29. When the analyzer finishes the test, repeat steps 22 through 25.
Reflection Tracking (Forward) € Calibration Coefficient
30. Press |
31. When the analyzer finishes the test, repeat steps 22 through 25.
Load Match (Reverse) Calibration Coefficient
32. Press (System) SERVICE |
33. When the analyzer finishes the test, repeat steps 22 through 25.
Transmission Tracking (Reverse) Calibration Coefficient
35. When the analyzer finishes the test, repeat steps 22 through 25.
System Verification and 47
Performance Tests
Second Full 2-Port Calibration —
36. Connect the equipment as shown in Figure 1-29. —
HP 8753E
NETWORK ANALYZER
PORT 1 ‘ ANALYZER
TEST PORT Z
CABLE 24 INCH _
ei REFERENCE
TEST PORT 1
sg622e
Figure 1-29. Second Full 2-Port Calibration Test Setup
37. Press (Preset) (Start) (30) (Zw). —
38. ;
39.
40. Connect the “open” end of the open/short combination (supplied in the calibration kit) to
reference test port 1.
41. Press R
42. Connect the “short” end of the open/short combination to reference test port 1.
44. Replace the open/short combination with the 50 ohm termination (supplied in the
calibration kit).
Wait for the message COMPUTING CAL COEFFICIENTS to disappear from the analyzer display
before proceeding to the next step. 00°
48 System Verification and
Performance Tests
23. Connect the equipment as shown in Figure 1-30.
HP 8753E
NETWORK ANALYZER
"a iin.
0000 0000
0
D
CABLE
24 INCH
sg6lBe
Figure 1-30. Transmission Calibration Test Setup
54. Press |
55. Press | .
Load Match (Forward) Calibration Coefficient
56. Press |
57. When the test is done, press MA
58. Using the front panel knob, locate the maximum value of the data trace for the 300 kHz to
1.3 GHz frequency range.
59. Write the maximum value on the “Performance Test Record.”
60. Repeat the previous three steps for the other frequency range(s) listed on the
“Performance Test Record.”
Directivity (Reverse) Calibration Coefficient
51. Press (Gyster) $
62. When the analyzer finishes the test, repeat steps 57 through 60.
Source Match (Reverse) Calibration Coefficient
55. Press 5 GD. At the prompt, press 3
64. When the analyzer finishes the test, repeat steps 57 through 60.
Reflection Tracking (Reverse) Calibration Coefficient
65. Press :
66. When the analyzer finishes the test, repeat steps 57 through 60.
System Verification and 43
Performance Tests
10. System Trace Noise (Only for Analyzers without
Option 006)
Frequency Range | Ratio
System Trace Noise
System Trace Noise
(Magnitude!) (Phase!)
30 kHz to 3 GHz | AR <0.006 dB rms <0.038” rms
30 kHz to 3 GHz | B/R <0.006 dB rms <(0.038° rms
1 At +5 dBm into test port, 3 kHz IF bandwidth, and CW sweep.
Required Equipment for 5001 Analyzers
Cable, APC-7, 24-inch
LEE
Required Equipment for 759 Analyzers
Cable, 750, Type-N 24-inch
Analyzer warmup time:
Perform this test to measure the system trace noise at a designated frequency in both the A/R
1 hour
and B/R ratioed measurements.
FF BF OF oa om % # # EF FF bow 3 BE OF ou ox ok EF OE OEE & KE 0 8 4K OE FOF OE FOR OE 4 a 4 x oA
1. Connect the equipment as shown in Figure 1-31.
4. Press (Marker Fctn) М
feature.
50 System Verification and
Performance Tests
HP
8753E
NETWORK ANALYZER
0000 0000
D
D
D
00
о
НР P/N 8120-4779
HP P/N 8120-2408
sgt 18e
N to activate the instrument’s statistic
System Trace Noise for A/R Magnitude
5. Press (Meas) -
7. When the analyzer displays the “Hid” annotation, press 1
8. Write the s.dev (standard deviation) value, which appears on the analyzer display, on the
“Performance Test Record.”
System Trace Noise for A/R Phase
9. Press (Format) PHASE
10. Press T
11. When the analyzer finishes the number of sweeps, press (Scale Ref) |
12. Write the s.dev value on the “Performance Test Record.”
System Trace Noise for B/R Magnitude
13. Press (Mess)
14. Press
15. When the analyzer finishes the number of sweeps, press AU
16. Write the s.dev value on the “Performance Test Record.”
System Trace Noise for B/R Phase
17. Press (Formar) PHASE
18. Press - ее)
19. When the analyzer finishes the number of sweeps, press |
20. Write the s.dev value on the “Performance Test Record.”
In Case of Difficulty
1. Perform the “ADC Offset Correction Constants” procedure, located in the “Adjustments and
Correction Constants” chapter.
2. Repeat this performance test.
3. Suspect the A10 Digital IF board assembly if the analyzer still fails the test.
System Verification and 51
Performance Tests
11. System Trace Noise (Only for Analyzers with Option 006)
Specifications
Frequency Range | Ratio | System Trace Noise | System Trace Noise
(Magnitude?) (Phase!)
| 30 kHz to 3 GHz | A/R <0.006 dB rms <0.038° rms
30 kHz to 3 GHz | B/R | <0.006 dB rms <0.038° rms
3 GHz to 6 GHz | A/R <0.010 dB rms <0.070° rms
3 GHz to 6 GHz | B/R <0.010 dB rms <0.070° rms
1 At +5 dBm into test port, 3 kHz IF bandwidth, and CW sweep.
Required Equipment
Cable, APC-7, 24-inch ................................... PS HP P/N 8120-4779
Analyzer warmup time: 1 hour
Perform this test to measure the system trace noise at designated CW frequencies in both the
A/R and B/R ratioed measurements.
1. Connect the equipment as shown in Figure 1-32.
HP 8753E
NETWORK ANALYZER
pa —
sgb18e
3. Press
feature.
| to activate the instrument's statistic
52 System Verification and
Performance Tests
System Trace Noise for A/R Magnitude from 30 kHz to 3 GHz
4. Press :
5 Press (Menu) CW F S JO.
6. When the analyzer finishes the number of sweeps, press (Scale Ref]
7. Write the s.dev value shown, which appears on the analyzer display, on the “Performance
Test Record.”
System Trace Noise for A/R Magnitude from 3 GHz to 6 GHz
8. Press (Mens) CW FREQ (© CD :
9. When the analyzer finishes the number of sweeps, press Al LE.
10. Write the s.dev value, which appears on the analyzer display, on the “Performance Test
Record.”
System Trace Noise for A/R Phase from 3 GHz to 6 GHz
12. press ;
13. When the analyzer finishes the number of sweeps, press (Scale Ref) A
14. Write the s.dev value, which appears on the analyzer display, on the “Performance Test
Record.”
System Trace Noi
15. Press (am) CW FREQ (3) (S/o) TH
16. When the analyzer finishes the number of sweeps, press Al
17. Write the s.dev value, which appears on the analyzer display, on the “Performance Test
Record.”
19. When the analyzer finishes the number of sweeps, press |
20. Write the s.dev value, which appears on the analyzer display, on the “Performance Test
Record.”
System Verification and 53
Performance Tests
System Trace Noise for B/R Magnitude from 3 GHz to 6 GHz
cw FREQ (6) E)
23. Write the s.dev value, which appears on the analyzer display, on the “Performance Test
Record.”
26. Write the s.dev value, which appears on the analyzer display, on the “Performance Test
Record.”
28. When the analyzer finishes the number of sweeps, press
29. Write the s.dev value, which appears on the analyzer display, on the “Performance Test
Record.”
In Case of Difficulty
1. Perform the “ADC Offset Correction Constants” procedure, located in the “Adjustments and
Correction Constants” chapter.
2. Repeat this performance test.
3. Suspect the A10 Digital IF board assembly if the analyzer still fails the test.
54 System Verification and
Performance Tests
12. Test Port Input Impedance
Specifications
Frequency Range | Test Port Input | Return Loss
300 kHz to 1.3 GHz Port 1 >18 dB
1.3 GHz to 3 GHz Port 1 >16 dB
3 GHz to 6 GHz Port 1 >14 dB
300 kHz to 1.3 GHz Port 2 >18 dB
1.3 GHz to 3 GHz Port 2 >16 dB
3 GHz to 6 GHz Port 2 >14 dB
Required Equipment for 500 Analyzers
Cable, APC-7, 24-inch ........................... SO .... HP P/N 8120-4779
Calibration Kit, 7-mm
Required Equipment for 751 Analyzers
Cable, 75Q, Type-N, 24-inch
Calibration Kit, 7502, Type-N
Analyzer warmup time: 1 hour
Perform this test to measure the return loss of each input test port.
& # & EF # #6 F & * = & ee ee ee eee yuu... HP 850318
PS LA Lea eee ess HP P/N 8120-2408
a Ce enicerec enana HP 85036B
System Verification and 55
Performance Tests
1. Connect the equipment as shown in Figure 1-33.
HP 8753E
NETWORK ANALYZER
at” zan
0000 0000
a
PORT 1 PORT 2
jt CABLE 24 INCH as
REFERENCE TEST
PORT 1 >
1 —
aaa, No aps aus
=, X
a
OPEN SHORT LOAD —
sgbz23e
Figure 1-33. S11 1-Port Cal Test Setup
2. Press (Preset) (Menu) ]
and select the appropriate calibration kit:
® [Г your analyzer is 509, press
® If your analyzer is 750, press
8. When the analyzer displays the prompt CONNECT STD THEN PRESS KEY TO MEASURE, connect
a short to reference test port 1.
. Press |
14. Connect the equipment as shown in Figure 1-34.
56 System Verification and
Performance Tests
15.
16.
. Connect the equipment as shown in Figure 1-35.
HP 8753E
NETWORK ANALYZER
QT 24-INCH CABLE
596 15е
Figure 1-34. Test Port 2 Input Impedance Test Setup
Press to turn the analyzer’s marker 1 on. Use the front panel knob to locate
the maximum value of the data trace for each of the frequency ranges listed in the
“Performance Test Record.”
Write these maximum values on the “Performance Test Record.”
HP 8753E
NETWORK ANALYZER
“a ias
—# CABLE 24 INCH
REFERENCE TEST |
PORT 2 LC
|
= Soot wy
ООС
OPEN SHORT LOAD
sgbZ4e
Figure 1-35. S22 1-Port Cal Test Setup
. Press (@D €
21.
22.
When the analyzer displays the prompt CONNECT STD THEN PRESS KEY TO MEASURE, connect
a short to reference test port 2.
Press R
System Verificationand 57
Performance Tests
23. At the prompt, connect a load to reference test port 2.
. Press I . >
27. Connect the equipment as shown in Figure 1-36.
HP 8753E
NE TWORK ANALYZER
ui Bh,
PORT 1 PORT 2
CABLE
24 INCH
$96 18е |
Figure 1-36. Test Port 1 Input Impedance Test Setup
28. Press to activate the analyzer’s marker 1. Use the front panel knob to locate
the maximum value of the data trace for each of the frequency ranges listed in the
“Performance Test Record.” —
29. Write the maximum values on the “Performance Test Record.”
In Case of Difficulty
1. Suspect the A10 digital IF board assembly if the analyzer fails BOTH test port tests.
2. Refer to the “Receiver Troubleshooting” chapter of the HP 8753E Service Guide for more
troubleshooting information.
b8 System Verification and
Performance Tests
Nn”
13. Test Port Receiver Magnitude Dynamic Accuracy
Specifications
HP8753E Magnitude Dynamic Accuracy 0.3 to 3000 MHz HP8753E Magnitude Dynamic Accuracy 3-8 GHz
Reference Power Level Reference Power Level
—[+-- -10dBm —[}+— -10dBm
_ 1 —>X— -2 dBm р 1 —>X — -20 dBm
8 —&— -30dBm g -—A— -30 dBm
© 7
a SS
3 0.1 3 0.1
< <
0.01 0.01
10 0 -10 -20 -30 -40 -50 -60 -70 -80 -90 -100 10 0 -10 -20 -30 -40 -50 -60 -70 -80 -90 -100
Test Port Power (dBm) Test Port Power (dBm)
Required Equipment
Power Meter LE EE EE EE EE EE EE EE EE EE EE EEE EET EE EE
Power Sensor мк жж % + 4 + 4 + #4 # u koa ow oa ow ow Po 3 y + = æ # + + + 44 4 # +4 4 # = + 5 % + + % + + + + # = + $ 4 x +
Step Attenuator, 110 dB ....... 1 1 aan LL L Lan
(See notes on the following page.)
Adapter (2), APC-7 to Type-N (£) 11 nas
Adapter, Type-N (f) to Type-N (f) i...
Cable (3), 500, Type-N, 24-inch ........................... В
Cable, HP-IB ........... Le Lea ae dan da aa a a nana aan
Diskette, 3.5inch .................... TT ER
Calibration Kit, Type-N, 50 Q............... eee
Additional Required Equipment for 750 Analyzers
Minimum Loss Pad (2), 50Qto 75Q ................ a a ana
Analyzer warmup time: 1 hour
Cea HP 436A/437B/438A
ea .. HP 8482A
HP 8496A Option 001, H18
Cee ... HP 11524A
Ce HP P/N 1250-0777
ea 8120-4781
Ce .. HP 10833A
VS Ce any
Ca HP 85032B
Cee HP 11852B
System Verification and 59
Performance Tests
Note
The HP 8496A step attenuator (Option 001, H18) comes with a special
calibration that supports the measurement uncertainties expressed in the test
record for this performance test.
The special calibration consists of two measurements. The first is a
measurement of the attenuation at each step. The data reported for this
measurement have the following uncertainites:
= +0.006 dB from 0 to 40 dB
m +0.015 dB from >40 to 80 dB
a +0.025 dB from >80 to 90 dB
m +0.05 dB >90 dB
The second calibration measurement characterizes match stability between
attenuator settings for each attenuator port. The vector difference of Si;
or (S22) between the reference attenuation step and all the other steps is
measured. The magnitude of this difference is certified to be <0.0316 (>30 dB).
Note
The HP 8496A used for this test will have known attenuator errors for
attenuations up to 100 dB using a test frequency of 30 MHz. The attenuation
used as a reference is 0 dB. If the available calibration data is not expressed
as attenuation errors, it can be converted to such a form by the following
equation:
(actual attenuation) — (expected attenuation) = attenuator error
Actual attenuation values that are greater than the expected attenuation
values will result in positive errors. Actual attenuation values that are less than
the expected attenuation values will result in negative errors.
60 System Verification and
Performance Tests
e
Initial Calculations
1. Fill in the attenuator error values (referenced to 0 dB attenuation) in Table 1-1 by referring
to the calibration data for the HP 8496A step-attenuator. Refer to the note on the previous
page.
a. Find the column in the HP 8496A attenuation error table that pertains to the
attenuation errors for 30 MHz.
b. Starting with the “10 dB” step in this column, write down the value in the
corresponding space in Table 1-1 for column “B.” This value should be placed in the row
for the 10 dB HP 8496A setting.
с. Continue transferring the remaining values of the HP 8496A attenuation errors to
column “B” in Table 1-1.
2. In Table 1-1, transfer the 10 dB error value located within the parenthesis in column “B” to
each space column “C.”
Table 1-1. Magnitude Dynamic Accuracy Calculations
A B C D E F
(B — C) (Е - D)
8496A Attn. Error 10 dB Error Value Attn. Error Expected Expected
Attn. (ref 0 dB) (ref 10 dB) Measurement Measurement
(dB) (dBm) (corrected)
(dBm)
0 0 dB er re a 10
10 oe) J 0 dB 0
20 ere —————]] rt — 10
30 I. — J, — 20
40 re re re BL — 30
50 re rm ee —— [ds — 40
60 paa aaa ————. — 50
70 man aaa aaa mamas | — 60
80 re — — ———— — 70
90 rm ——] era — 80
. The values in column “D” result from changing the reference attenuation of the calibration
data of the HP 8496A to 10 dB.
Calculate the attenuation error values for this column by subtracting the values in column
“C” from the values in column “B” (B — C = D).
. The values in column “F” result from correcting the expected measurement value by the
amount of attenuator error.
Calculate the values in this column by subtracting the values in column “D” from the
values in column “E” (E — D = F).
5. Transfer the values from column “F” in Table 1-1 to column “F” in the “Performance Test
Record” for both test ports.
System Verification and 61
Performance Tests
Port 1 Power Meter Calibration
6. Zero and calibrate the power meter. (Refer to the power meter manual for details on this
procedure.)
7. Connect the equipment as shown in Figure 1-37.
*
ADAPTER
500/750
MINIMUM LOSS PAD
CABLE 500
TYPE-N
24 INCH
110 dB
STEP
ATTENUATOR
+
ADAPTER
APC—7 TO
ТУРЕ-М (f)
CABLE SON
TYPE-N
24 INCH
110 dB
STEP
ATTENUATOR
* DIRECT CONNECT ON
Figure 1-37. Power Meter Calibration for Magnitude Dynamic Accuracy
62 System Verification and
Performance Tests
HP 8753E
NETWORK ANALYZER
WITH OPTION 075
POWER METER
ro
INPUT 509
ADAPTER
500/750
MINIMUM LOSS PAD
POWER SENSOR
Oo | ’
*
HP 8753E
NETWORK ANALYZER
a.
5958
Be |
POWER METER
ob
o a
50 o
=)
PORT 1
INPUT 5010
ADAPTER
APC-7 TO
ТУРЕ-М (К) |
POWER SENSOR
o ’
8. Set the HP 8496A to 10 dB.
9. Set the following analyzer parameters:
10. Set up the HP 8753E for power meter calibration:
a. Select the HP 8753E as the system controller:
b. Set the power meter’s address:
C. Select the appropriate power meter by pressing ] IWER
model number is displayed (HP 436A or HP 438A/437).
d. Select the cal kit and enter the power sensor calibration data.
until the correct
. (enter the power
11. Take a power meter calibration sweep.
12. Verify that the power meter reads approximately —20 dBm.
System Verification and 63
Performance Tests
Adapter Removal Calibration
13. Connect the equipment as shown in the Figure 1-38:
HP 8753E
NETWORK ANALYZER
WITH OPTION 075
al Sa 5000
flee Eas
o BC
Sd ge 86 <
= oO
Ф Ф
PORT 2
ADAPTER
509/750
MINIMUM LOSS PAD
MINIMUM LOSS PAD CABLE 50
TYPE-N
CABLE 500 24 INCH
ТУРЕ-М (т) ТО (т)
24 INCH
(т) ТО (т) ADAPTER
TYPE-N (f)
TO TYPE-N (1)
HP B753t
NETWORK ANALYZER
\
/
ADAPTER
АРС-7 ТО
TYPE-N (f)
ADAPTER
АРС-7 ТО
ТУРЕ-М (К)
CABLE Son
TYPE-N
24 INCH
CABLE 500
(m) TO (m)
TYPE—N
24 INCH
(т) ТО (т) ADAPTER
TYPE-N (f)
TO TYPE-N (f)
* DIRECT CONNECT ON
sg6118e |
Figure 1-38. Full 2-Port Calibration with Adapter Removal
14. Perform a full 2-port error correction with isolation.
Note When you are performing error-correction for a system that has type-N test
port connectors, the softkey menus label the sex of the test port connector
— not the calibration standard connector. For example, the label $
refers to the short that will be connected to the female test port.
15. Save the results to disk. Name the file “PORT1.”
16. Move the adapter to reference test port 1 and perform another full 2-port error correction.
17. Save the results to disk. Name the file “PORT2.”
64 System Verification and
Performance Tests
24.
Save the results of the new cal set.
System Verification and 65
Performance Tests
Measure Test Port 2 Magnitude Dynamic Accuracy —
25. Remove the Type-N (1) to (f) adapter and connect the equipment as shown in Figure 1-39. —
Confirm that the step attenuator is set to 10 dB.
HP 8753E
NETWORK ANALYZER —
WITH OPTION 075
hh,
—
co 0000
0000
ADAPTER [ [MINIMUM LOSS PAD
500/750 | |
MINIMUM LOSS PAD | | CABLE 509
T TYPE-N
24 INCH
CABLE 500
TYPE—N
24 INCH
110 dB
STEP ATTENUATOR
HP 8753E _
NETWORK ANALYZER
gual, = SN
al ao DODO
a
C 000000 is
| 4 888 © —
= giao o
® @
PORT 2 a
ADAPTER
АРС-7 TO
ADAPTER ТУРЕ-М (СР) —
APC—7 TG
TYPE-N (f) CABLE 50%
TYPE-N
CABLE 500 24 INCH
TYPE-N nó
24 INCH
110 dB
STEP ATTENUATOR
+ DIRECT CONNECTION Lies
sabôle
Figure 1-39. Magnitude Dynamic Accuracy Measurement
66 System Verification and
Performance Tests
26. To set up the dynamic accuracy measurement, press the following:
(Marker Fctn)
— 27. Wait for the sweep to finish, then press 1
28. Set the step attenuator to 0 dB.
— 29. Press (Menu)
30. Write the mean value (which appears on the analyzer's display) in the “Test Port
— Measurement” column of the “Performance Test Record.” This column is also labeled “G.”
31. Repeat steps 28 through 30 for each setting of the step attenuator.
— 32. Calculate dynamic accuracy for each step by using the formula |G — F|. Place these values
in the appropriate column of the “Performance Test Record.”
System Verification and 67
Performance Tests
Measure Test Port 1 Magnitude Dynamic Accuracy
33.
34.
35.
36.
37.
38.
39.
40.
Set the step attenuator to 10 dB.
Wait for the sweep to finish, then press I
Set the step attenuator to 0 dB.
Press |
Write the mean value (which appears on the analyzer’s display) in the “Test Port
Measurement” column of the “Performance Test Record.” This column is also labeled “G.”
Repeat steps 36 through 38 for each setting of the step attenuator.
Calculate dynamic accuracy for each step by using the formula |G — F|. Place these values
in the appropriate column of the “Performance Test Record.”
In Case of Difficulty
1.
If the analyzer fails the test at ALL power levels, be sure you followed the recommended
attenuator settings as listed in the “Performance Test Record.” Repeat this performance test.
. If both test port measured values are out of specifications:
a. Recalibrate the power meter.
b. Repeat this performance test.
. If the analyzer fails either test port 2 or test port 1 dynamic accuracy at lower power levels:
a. Perform the “IF Amplifier Correction Constants” and “ADC Offset Correction Constants”
procedures (located in the “Adjustments and Correction Constants” chapter).
b. Repeat this performance test.
c. If it still fails, replace the A10 Digital IF assembly.
d. Repeat the two adjustment procedures mentioned in this step and then repeat this
performance test.
System Verification and
Performance Tests
14. Test Port Receiver Magnitude Compression
Specifications
Frequency Range | Test Port Magnitude! |
300 kHz to 3 GHz | Port 1 <0.45 dB
| 3 GHz to 6 GHZ? Port 1 <0.80 dB
300 kHz to 3 GHz | Port 2 <0.45 dB
3 GHz to 6 GHz? | Port 2 <0.80 dB
1 With a 10 Hz IF bandwidth.
2 Only for analyzers with Option 006 — 30 kHz to
6 GHz range.
Required Equipment for 500 Analyzers
Cable, APC-7, 24-inch .......... aaa reaeroa, . HP P/N 8120-4779
Required Equipment for 750 Analyzers
Cable, 750, Type-N, 24-inch ......... a Aa aa aa dan a na aan a ee nana о HP P/N 8120-2408
Analyzer warmup time: 1 hour
Perform this test to verify the compression/expansion magnitude levels of the analyzer’s test
port receiver samplers.
Test Port 2 Magnitude Compression
1. Connect the equipment as shown in Figure 1-40.
HP 8753E
NE TWORK ANALYZER
sgb18e
Figure 1-40. Test Port Magnitude Compression Test Setup
2. Press 3
3 Press (Aug) IF BW (19) (D.
4. Press (Menu) |
5. Press
System Verification and 69
Performance Tests
10.
11.
12.
13.
14.
15.
16.
17.
18.
Press A
Write the absolute value of marker 2 in the “Performance Test Record.”
Repeat steps 12 through 17 for the other frequencies listed for Port 2 on the “Performance
Test Record.”
Test Port 1 Magnitude Compression
- Press
25.
26.
70
Write the absolute value of the marker 2 reading in the “Measured Value” column of the
“Performance Test Record.”
Repeat steps 19 through 25 for the other CW frequencies listed for Port 1 in the
“Performance Test Record.”
System Verification and
Performance Tests
Nr’
In Case of Difficulty
1. If the analyzer fails “Test Port 2 Magnitude Compression”:
a. Repeat this test.
b. Replace the A6 B sampler assembly if the analyzer still fails the test.
2. If the analyzer fails “Test Port 1 Magnitude Compression”:
a. Repeat this test.
b. Replace the A5 A sampler assembly if the analyzer still fails the test.
System Verification and 71
Performance Tests
15. Test Port Receiver Phase Compression
Specifications
CW Frequency | Test Port | Phase!
300 kHz to 3 GHz| Port 1 <6°
3 GHz to 6 GHz? | Port 1 | <7.5°
300 kHz to 3 GHz| Port 2 <6°
3 GHz to 6 GHz? | Port 2 | <7.5°
1 With 10 Hz IF bandwidth.
2 Only for analyzer with Option 006 — 30 kHz
to 6 GHz range.
Required Equipment for 501 Analyzers
Cable, APC-7, 24-inch .......... RS В ВНИИ .... HP P/N 8120-4779 __
Required Equipment for 750 Analyzers
Cable, 750, Type-N, 24-inch ...... ..... aa een ee ea ann a an HP P/N 8120-2408 —
Analyzer warmup time: 1 hour
Perform this test to verify the compression/expansion phase relationships of the analyzer's test
port receiver samplers.
Test Port 2 Phase Compression — —
1. Connect the equipment as shown in Figure 1-41.
HP 8753E
NE TWORK ANALYZER
0000 0000
D
D
ao
| ==)
PORT 1 PORT 2
CABLE
24 INCH
sg618e
Figure 1-41, Test Port Phase Compression Test Setup —
2. Press (Preset) (Mess) |
эре)
* Press (Menu) £
12 System Verification and
Performance Tests
— 6. Press (Menu)
~— 7. At the end of the sweep, press 4
8. Press (Marker Fctn) y
— 9. Press X
10.
(Marker Fctn) N.
— 11. Write the absolute value of the marker 2 reading in the “Measured Value” column of the
“Performance Test Record.”
_ 12. Repeat steps 5 to 11 for the other CW frequencies listed for Port 2 in the “Performance Test
Record.”
Test Port 1 Phase Compression
13. Press (Meas) я
14. Press (Menu) CW FREQ (ES
_— 15. press |
16.
17. Press (Marker)
o 18. Press
19. Write the absolute value of the marker 2 reading in the “Measured Value” column of the
“Performance Test Record.”
20. Repeat steps 14 to 19 for the other CW frequencies listed for Port 1 in the “Performance
Test Record.”
In Case of Difficulty
— 1. If the analyzer fails the “Test Port 2 Phase Compression” test:
a. Repeat this test.
— b. Replace the A6 B sampler assembly if analyzer still fails the test.
2. If the analyzer fails the “Test Port 1 Phase Compression” test:
— a. Repeat this test.
b. Replace the A5 A sampler assembly if analyzer still fails the test.
System Verification and 73
Performance Tests
16. Test Port Output/Input Harmonics (Option 002 Analyzers
without Option 006 Only)
Specifications
Test Port | Harmonic Limit
Output 2nd < —25 @Вс @ +10 dBm
Output 3rd < —25 dBc @ +10 dBm
Input Port 1 2nd < —15 dBc @ +8 dBm
Input Port 1 3rd < -30 dBc @ +8 dBm
Input Port 2 2nd < —15 @Вс @ +8 dBm
Input Port 2 3rd < 30 dBc @ +8 dBm
Equipment Required for 501) Analyzers
Cable, APC-7, 24-inch .............. Ce HP P/N 8120-4779
Attenuator (2), 20 dB, APC-7 ...... aaa HP 8492A Option 020
Equipment Required for 759 Analyzers
Minimum Loss Pad (2) ........o 0 aa aa a en aereo. HP 11852B
Cable, Type-N ............ LL LA AAA AAA AAA A A a ed aa a aan LL ee aa a das an HP P/N 8120-2408
Attenuator (2), 20 dB, Type-N ............. 000 e HP 8491A Option 020
Analyzer warmup time: 30 minutes
Perform this test to determine the spectral purity of the HP 8753E input and output test ports.
Note The test port input 3rd harmonic specifications are better than the test port
output 3rd harmonic specifications.
Test Port Output Worst Case 2nd Harmonic
1. Press
2. Press to set the frequency range.
3. Press (Avg) IF BW to set the IF bandwidth to 10 Hz.
4. Connect the equipment as shown in Figure 1-42.
14 System Verification and
Performance Tests
- Press
+ After one sweep, press
Press (System)
- After one sweep, press
- Press (Marker Fctn)
10.
HP 8753E
| HP B753E NETWORK ANALYZER
NETWORK ANALYZER WITH OPTION 075
na
Fals Os
1 ooo
| Ho
i 15:00 C6
E ро | ее o ©
Ф Ф = Ф
PORT 1 PORT 2
_ ADAPTER ADAPTER
seen 500/750 500/750
20 dB FIXED MINIMUM LOSS PAD MINIMUM LOSS PAD
ATTENUATORS
- 20 dB FIXED CABLE ТУРЕ-М
ATTENUATORS
sgb27e
Figure 1-42. Test Port Output Harmonics Test Setup
: to normalize the trace.
“to get a better viewing of the trace.
Write the marker 1 value (which appears on the analyzer display) on the “Performance Test
Record.” This is the worst case test port output 2nd harmonic.
Test Port Output Worst Case 3rd Harmonic
11.
12.
13.
14.
15.
16.
17.
18.
Press to change the stop freq
Press HARMO
After one sweep, press \Y DATA/MEM
Press A (x1) to get a better viewing of the trace.
Press HAR] MC =
After one sweep, press (Scale Ref)
Press (Marker Fctn) Mi
Write the marker 1 value on the “Performance Test Record.”
System Verificationand 75
Performance Tests
Port 1 Input Worst Case 2nd Harmonic
19. Connect the equipment as shown in Figure 1-43.
HP 8753E
HP 8753E NETWORK ANALYZER
NETWORK ANALYZER WITH OPTION 075
a —
a! oo ADO
Imes
S| 8. ong o
а Оба
= 5166 о
Ф Ф =
er чо
PORT 1 PORT 2 PORT 1 PORT 2
ADAPTER ADAPTER
500/750 500/750
MINIMUM LOSS PAD MINIMUM LOSS PAD
CABLE APC—7 CABLE ТУРЕ-М
Ne
59628е
Figure 1-43. Receiver Harmonics Test Setup
24 Grp f to normalize the trace.
(Scale Ref)
20. After one sweep, press to get a better viewing of the trace.
27. Press (Marker Fen) №
28. Write the marker 1 value (which appears on the analyzer display) on the “Performance Test
Record.” This is the worst case port 1 input (receiver channel A) 2nd harmonic.
Port 1 Input Worst Case 3rd Harmonic
29. Press (Stop) = D (G/n) to change the stop frequency for measuring the receiver 3rd harmonic.
30. Press (System) |
31. After one sweep, press .
32. Press (Scale Ref) See RD) !
33. Press (System) |
34. After one sweep, press | -
i to normalize the trace.
to get a better viewing of the trace.
76 System Verification and
Performance Tests
ance Test Record.”
Port 2 Input Worst Case 2nd Harmonic
38
39
40.
41.
42.
43.
44.
Press to set the stop frequency
Press 1 SI À 2 В.
After one sweep, press I to normalize the trace.
Press (Exam) н ee
After one sweep, press A to get a better viewing of the trace.
Press (Marker Fetn) M
Write the marker 1 value (which appears on the analyzer display) on the “Performance Test
Record.” This is the worst case port 2 input (receiver channel B) 2nd harmonic.
Port 2 Input Worst Case 3rd Harmonic
45.
46.
47.
48.
49.
50.
51.
52.
Press to change the stop frequency for measuring the receiver 3rd harmonic.
Press HARMON
After one sweep, press (Display) DATA—MEMORY DATA/MEM “to normalize the trace.
Press à
Press (System) HARMONIC MEAS
After one sweep, press
Press MAN
Write the marker 1 value on the “Performance Test Record.”
; to get a better viewing of the trace.
System Verification and 77
Performance Tests
17. Test Port Output/Input Harmonics (Option 002 Analyzers
with Option 006 Only)
Specifications
Test Port | Harmonic Limit
Output 2nd < —25 аВс @ +10 dBm
Output 3rd < —25 dBc @ +10 dBm
Input Port 1 2nd < —15 dBc @ +8 dBm
Input Port 1 3rd < —30 dBce Q +8 dBm
Input Port 2 2nd < —15 dBc @ +8 dBm
Input Port 2 3rd < —30 dBc @ +8 dBm
Equipment Required
Cable, APC-7, 24-inch ............... aaa. aaa HP P/N 8120-4779
Attenuator (2), 20 dB ................. ia ainia a HP 8492A Opt 020
Analyzer warmup time: 30 minutes
Perform this test to determine the spectral purity of the HP 8753E input and output test ports.
Note The test port input 3rd harmonic specifications are better than the test port
output 3rd harmonic specifications.
78 System Verification and
Performance Tests
ld aia
Test Port Output Worst Case 2nd Harmonic
1.
Press F
to set the test port power to +10 dBm.
2. Press (м/и) to set the frequency range.
to set the IF bandwidth to 10 Hz.
4. Connect the equipment as shown in Figure 1-44.
10.
- Press (ear) :
HP 8753E
NETWORK ANALYZER
a” № v00
al oo 9000
9 ooo 0000
9 966 000000
5 OO 0000900
| 5 ob 868 ©
= 8190 o
P za
“E ce
PORT 1 PORT 2
CABLE APC—7
20 dB FIXED 24 INCH
ATTENUATORS
549629е
Figure 1-44. Test Port Output Harmonics Test Setup
After one sweep, press : 4 to normalize the trace.
After one sweep, press (Scale Ref} | to get a better viewing of the trace.
Write the marker 1 value (which appears on the analyzer display) on the “Performance Test
Record.” This is the worst case test port output 2nd harmonic.
System Verificationand 79
Performance Tests
Test Port Output Worst Case 3rd Harmonic
11.
12.
13.
14.
16.
17.
18.
Press (G/n) to change the stop frequency to 1 GHz.
M to normalize the trace.
Press (Scale Ref) - to get a better viewing of the trace.
Press HARMONIC MEAS HARMONIC THIRD
After one sweep, press {Scale Ref)
Press (Marker Fctn | M
Write the marker 1 value on the “Performance Test Record.”
System Verification and
Performance Tests
Port 1 Input Worst Case 2nd Harmonic
Na 19. Connect the equipment as shown in Figure 1-45.
HP 8753E
NETWORK ANALYZER
CABLE APC—7
24 INCH
sg630e
Figure 1-45. Receiver Harmonics Test Setup
24.
25.
26.
27.
28.
After one sweep, press M to normalize the trace.
Press Hi
After one sweep, press (Scale Ref)
Press (Marker Fctn)
Write the marker 1 value (which appears on the analyzer display) on the “Performance Test
Record.” This is the worst case port 1 input (receiver channel A) 2nd harmonic.
to get a better viewing of the trace.
System Verification and 81
Performance Tests
Port 1 Input Worst Case 3rd Harmonic
29. Press to change the stop frequency for measuring the receiver 3rd harmonic.
30. Press ]
31. After one sweep, press ] \/MEM to normalize the trace.
32. Press | 2 to get a better viewing of the trace.
33. Press 1 ).
34. After one sweep, press
35. Press I
36. Write the mar
37. I
Port 2 Input Worst Case 2nd Harmonic
38. Press
39. Press (Hex)
40 \ A to normalize the trace.
42.
to get a better viewing of the trace.
43.
44. Write the marker 1 value (which appears on the analyzer display) on the “Performance Test
Record.” This is the worst case port 2 input (receiver channel B) 2nd harmonic.
Port 2 Input Worst Case 3rd Harmonic
45. Press to change the stop frequency for measuring the receiver 3rd harmonic.
46. Press №
47. After one sweep, press (Display) { to normalize the trace.
19. Press Gruem) E
50. After one sweep, press (Scale Ref)
Ol. Press (Marker Fcen) MI
to get a better viewing of the trace.
52. Write the marker 1 value on the “Performance Test Record.”
82 System Verification and
Performance Tests
_ Performance Test Record
For Analyzers with a Frequency Range of
30 kHz to 3 GHz
— Note See the next “Performance Test Record” section if your analyzer frequency
range is from 30 kHz to 6 GHz (Option 006).
Performance Test Record 1
HP 8753E Performance Test Record (1 of 12)
Calibration Lab Address:
Model HP 8753E
Serial No.
Firmware Revision
Description
Frequency Counter
Power Meter
Power Sensor
Calibration Kit
Verification Kit
Notes/Comments:
Ambient Temperature
Test Equipment Used:
Model Number
Report Number
Date
Last Calibration Date
Customer's Name
Performed by
Option(s)
Relative Humidity %
Trace Number Cal Due Date
2 Performance Test Record
HP 8753E Performance Test Record (2 of 12)
For 30 kHz—3 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E
Serial Number
Report Number
Date
PP 1. Test Port Output Frequency Range and Accuracy
CW Frequencies Min. Results Measured Max. Measurement
(MHz) (MHz) (MHz) (MHz) Uncertainty
(MHz)
0.03 0.029 999 7 0.030 000 3 + 0.000 000 050
0.3 0.299 997 0.300 003 + 0.000 000 520
5.0 4.999 950 5.000 050 + 0.000 009
16.0 15.999 840 16.000 160 + 0.000 028
31.0 30.999 690 31.000 310 + 0.000 054
60.999 999 60.999 390 61.000 610 + 0.000 105
121.0 120.998 790 121.001 210 + 0.000 207
180.0 179.998 200 180.001 800 + 0.000 307
310.0 309.995 900 310.003 100 + 0.000 528
700.0 699.930 000 700.007 000 + 0.001 192
1 300.0 1 299.987 1 300.013 + 0.002 212
2 000.0 1 999.980 2 000.020 + 0.003 403
3 000.0 2 899.970 3 000.030 + 0.005 104
Pb 2. External Source Mode Frequency Range
Test Frequencies (GHz) Results
0.010
0.020
0.100
1.000
2.000
3.000
Performance Test Record 3
HP 8753E Performance Test Record (3 of 12)
For 30 kHz—3 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E
Serial Number
Report Number
Date
bb 3. Test Port Output Power Accuracy
Test Frequencies Test Port Specification Measured Value Measurement
Output Power (dB) (dB) Uncertainty
(dBm) (dB)
Center Frequency
300 kHz 0 + 1 + 0.465
20 MHz 0 +1 + 0.10
50 MHz 0 +1 + 0.10
100 MHz 0 + 1 + 0.10
200 MHz 0 + 1 + 0.10
500 MHz 0 + 1 + 0.10
1 GHz 0 +1 + 0.13
2 GHz 0 + 1 + 0.13
3 GHz 0 + 1 + 0.27
Pb 4. Test Port Output Power Range and Linearity
Test Settings Results Power Level Specification Measurement
Measured Linearity (dB) Uncertainty
(dB) (dB) (dB)
CW Frequency = 300 kHz
- 15 + 0.2 + 0.03
— 13 + 0.2 + 0.03
— li + 0.2 + 0.03
— 9 + 0.2 + 0.02
- 7 + 0.2 + 0.02
— 5 + 0.2 + 0.02
— 3 + 0.2 + 0.02
— 1 + 0.2 + 0.02
1 + 0.2 + 0.03
+ 3 + 0.2 + 0.03
5 + 0.5 + 0.03
4 Performance Test Record
HP 8753E Performance Test Record (4 of 12)
For 30 kHz—3 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E
Serial Number
Report Number
Date
pb 4. Test Port Output Power Range and Linearity (continued)
Test Settings Results Power Level Specification Measurement
Measured Linearity (dB) Uncertainty
(dB) (dB) (dB)
+ 7 + 0.5 + 0.03
+ 9 + 0.5 + 0.08
+ 10 + 0.5 + 0.03
CW Frequency = 3 GHz
— 15 + 0.2 + 0.03
— 13 + 0.2 + 0.03
— 11 + 0.2 + 0.03
— 9 + 0.2 + 0.02
— 7 + 0.2 + 0.02
— 5 + 0.2 + 0.02
— 3 + 0,2 + 0.02
— 1 + 0.2 + 0.02
+ 1 + 0.2 + 0.03
+ 3 + 0.2 + 0.03
+ 5 + 0.5 + 0.03
+ 7 + 0.5 + 0.03
+ 9 + 0.5 + 0.03
+ 10 + 0.5 + 0.03
Performance Test Record 5
HP 8753E Performance Test Record (5 of 12)
For 30 kHz—3 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E
Serial Number
Report Number
Date
PP» 5. Minimum R Channel Level
CW Frequency Specification Test Port Power Measurement
(dB) Uncertainty
(dB)
300 kHz < —35 + 1.0
3.29 MHz < —35 + 1.0
3.31 MHz < —35 + 1.0
15.90 MHz < —35 +1.0
16.10 MHz < —35 + 1.0
30.90 MHz < —35 + 1.0
31.10 MHz < -35 + 1.0
1.6069 GHZ < —35 + 1.0
1.6071 GHz < —35 + 1.0
3.000 GHz < —35 + 1.0
bb 6. Test Port Input Noise Floor Level
Frequency Range Test Port IF Bandwidth | Specification Calculated Measurement
(dBm) Value Uncertainty
300 kHz - 3 GHz Port 1 d kHz — 82 N/A
300 kHz - 3 GHz Port 1 10 Hz — 102 N/A
300 kHz - 3 GHz Port 2 10 Hz - 102 N/A
300 kHz - 3 GHz Port 2 3 kHz — 82 N/A
6 Performance Test Record
HP 8753E Performance Test Record (6 of 12)
For 30 kHz—3 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E
Serial Number
Report Number
Date
PP 7. Test Port Input Frequency Response
Measured Value
Frequency Range Test Port Specification Measurement
(dB) (dB) Uncertainty
(dB)
300 kHz-—3 GHz Port 2 +1 0.47
300 kHz—3 GHz Port 1 +1 0.47
Pp 8. Test Port Crosstalk
Test Settings Specification Measured Value Measurement
(dB) (dB) Uncertainty
Crosstalk to Test Port 2
300 kHz—3 GHz < — 100 N/A
Crosstalk to Test Port 1
300 kHz—3 GHz < — 100 N/A
Performance Test Record 7
HP 8753E Performance Test Record (7 of 12)
For 30 kHz—3 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E Report Number
Serial Number Date
>») 9. Calibration Coefficients
Test Description Frequency Range Spec. Measured Value Measurement
(dB) (dB) Uncertainty
(dB)
Forward Direction
Directivity 300 kHz - 1.3 GHz > 35 + 0.9
Directivity 1.3 GHZ - 3 GHZ > 30 + 0.8
Forward Direction
Source Match 300 kHz - 1.3 GHz > 16 + 0.2
Source Match 1.3 GHz - 3 GHZ > 16 + 0.2
Forward Direction
Trans. Tracking 300 kHz - 1.3 GHz + 1.5 + 0.006
Trans. Tracking 1.3 GHz - 3 GHz + 1.5 + 0.009
Forward Direction
Refl. Tracking 300 kHz - 1.3 GHz + 1.5 + 0.001
Refl. Tracking 1.3 GHz - 3 GHz + 1.5 + 0.005
Reverse Direction
Load Match 300 kHz - 1.3 GHz > 18 + 0.1
Load Match 1.3 GHz - 3 GHZ > 16 + 0.2
8 Performance Test Record
HP 8753E Performance Test Record (8 of 12)
For 30 kHz—3 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E
Serial Number .
Report Number
Date
bb 9. Calibration Coefficients (continued)
Test Description Frequency Range Spec. Measured Value Measurement
(dB) (dB) Uncertainty
(dB)
Reverse Direction
Trans. Tracking 300 kHz - 1.3 GHz + 1.5 + 0.006
Trans. Tracking 1.3 GHz - 3 GHz + 1.5 + 0.009
Forward Direction
Load Match 300 kHz - 1.3 GHz > 18 + 0.1
Load Match 1.3 GHZ - 3 GHZ > 16 + 0.2
Reverse Direction
Directivity 300 kHz - 1.3 GHz > 35 + 0.9
Directivity 1.3 GHz - 3 GHZ > 30 + 0.8
Reverse Direction
Source Match 300 kHz - 1.3 GHz > 16 + 0.2
Source Match 1.3 GHz - 3 GHz > 16 + 0.2
Reverse Direction
Refl. Tracking 300 kHz - 1.3 GHz + 1.5 + 0.001
Refl. Tracking 1.3GHz - 3 GHz + 1.5 + 0.005
Performance Test Record 9
HP 8753E Performance Test Record (9 of 12)
For 30 kHz—3 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E
Serial Number
Report Number
Date
bb 10. System Trace Noise
CW Frequency Ratio Specification Measured Value Measurement
(GHZ) Uncertainty
3 A/R < 0.006 dB rms +0.001 dB
3 A/R < 0.038° rms +0.01°
3 B/R < 0.006 dB rms +0.001 dB
3 B/R < 0.038” rms +0.01°
PP 12. Test Port Input Impedance
Frequency Range Test Port Return Loss Specification Measurement
(dB) (dB) Uncertainty
(dB)
300 kHz—1.3 GHZ Port 2 > 18 + 1.5
1.3 GHz—3 GHz Port 2 > 16 + 1.5
300 kHz—1.3 GHz Port 1 > 18 + 1.5
1.3 GHz—3 GHz Port 1 > 16 + 1.5
10 Performance Test Record
HP 8753E Performance Test Record (10 of 12)
For 30 kHz—3 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E
Report Number
Serial Number Date
PP 13. Test Port Receiver Magnitude Dynamic Accuracy
G F |G — Fl
Test Port 8496A Test Port Expected Dynamic Spec. Meas.
Input Power Attn. Measurement Measurement Accuracy (dB) Uncer.
(dB) (dBm) (corrected) (Calculated) (dB)
(dBm)
Test Port 2
— 10 0 < 0.033 + 0.008
— 20 (Ref) 10 Reference Reference Reference | Reference Reference
— 30 20 < 0.031 + 0.008
— 40 30 < 0.042 + 0.008
— 50 40 < 0.057 + 0.008
— 60 50 < 0.098 + 0.017
— 70 60 < 0.247 + 0.017
— 80 70 < 0.725 + 0.017
— 90 80 < 2.097 + 0.017
— 100 90 < 5.399 + 0.027
Test Port 1
— 10 0 | < 0.033 + 0.008
— 20 (Ref) 10 Reference Reference Reference Reference Reference
— 30 20 < 0.031 + 0.008
— 40 30 < 0.042 + 0.008
— 50 40 < 0.057 + 0.008
— 60 50 < 0.098 + 0.017
— T0 60 < 0.247 + 0.017
— 80 70 < 0.725 + 0.017
— 90 80 < 2.097 + 0.017
— 100 90 < 5.399 + 0.027
Performance Test Record
11
HP 8753E Performance Test Record (11 of 12)
For 30 kHz—3 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E
Serial Number
Report Number
Date
PP 14. Test Port Receiver Magnitude Compression
Measurement
CW Frequency Test Port Measured Value Specification
(dB) (dB) Uncertainty
50 MHz Port 2 < 0.45 N/A
1 GHz Port 2 < 0.45 N/A
2 GHz Port 2 < 0.45 N/A
3 GHz Port 2 < 0.45 N/A
50 MHz Port 1 < 0.45 N/A
1 GHz Port 1 < 0.45 N/A
2 GHz Port 1 < 0.45 N/A
3 GHz Port 1 < 0.45 N/A
PP 15. Test Port Receiver Phase Compression
CW Frequency Test Port Measured Value Specification Measurement
(degrees) (degrees) Uncertainty
50 MHz Port 2 < 6° N/A
1 GHz Port 2 < 6° N/A
2 GHZ Port 2 < 6° N/A
3 GHz Port 2 < 6° N/A
50 MHz Port 1 < 6° N/A
1 GHz Port 1 < 6° N/A
2 GHz Port 1 < 6° N/A
3 GHz Port 1 < 6° N/A
12 Performance Test Record
HP 8753E Performance Test Record (12 of 12)
For 30 kHz—3 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E
Serial Number
Report Number
Date
№» 16. Test Port Output/Input Harmonics (Option 002 without Option 006)
Test Description Specification Measurement Value Measurement Uncertainty
(dBc) (dBc) (dB)
Test Port Output Harmonics
2nd < 25 + 1.5
3rd < 25 + 1.5
Port 1 Input Harmonics
2nd < 15 + 1.5
3rd < 30 + 1.5
Port 2 Input Harmonics
2nd < 15 + 1.5
3rd < 30 + 1.5
Performance Test Record 13
Performance Test Record
For Analyzers with a Frequency Range of
30 kHz to 6 GHz
Note
See the previous “Performance Test Record” section if your analyzer frequency
range is from 30 kHz to 3 GHz.
Performance Test Record
1
HP 8753E Performance Test Record (1 of 14)
Calibration Lab Address:
Serial No.
Model HP 8753E Option 006
Firmware Revision
Description
Frequency Counter
Power Meter
Power Sensor
Calibration Kit
Verification Kit
Notes/Comments:
Ambient Temperature
Test Equipment Used:
Model Number
Report Number
Date
Last Calibration Date
Customer’s Name
Performed by
Option(s)
Relative Humidity
Trace Number Cal Due Date
%
2 Performance Test Record
HP 8753E Performance Test Record (2 of 14)
For 30 kHz—6 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E Option 006
Serial Number
Report Number
Date
PP 1. Test Port Output Frequency Range and Accuracy
Test Frequencies Min. Results Measured Max. Measurement
(MHz) (MHz) (MHz) (MHz) Uncertainty
(MHz)
0.03 0.029 999 7 0.030 000 3 + 0.000 000 050
0.3 0.299 997 0.300 003 + 0.000 000 520
5.0 4.999 950 5.000 050 + 0.000 009
16.0 15.999 840 16.000 160 + 0.000 028
31.0 30.999 690 31.000 310 + 0.000 054
60.999 999 60.999 390 61.000 610 + 0.000 105
121.0 120.998 790 121.001 210 + 0.000 207
180.0 179.998 200 180.001 800 + 0.000 307
310.0 309.995 900 310.003 100 | + 0.000 528
700.0 699.930 000 700.007 000 + 0.001 192
1 300.0 1 209.987 1 300.013 + 0.002 212
2 000.0 1 999.980 2 000.020 + 0.003 403
3 000.0 2 999.970 3 000.030 + 0.005 104
4.0 3.999 960 4.000 040 + 0.006 8056
5.0 4.999 950 5.000 050 + 0.008 506
6.0 5.999 940 6.000 060 + 0.010 207
Performance Test Record 3
HP 8753E Performance Test Record (3 of 14)
For 30 kHz—6 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E Option 006
Serial Number
Keport Number
Date
bb 2. External Source Mode Frequency Range
Test Frequencies (GHz)
Result
0.010
0.020
0.100
1.000
2.000
3.000
4.000
5.000
6.000
Pp 3. Test Port Output Power Accuracy
Test Frequency Test Port Specification Measured Value (dB) Measurement
Output Power (dB) Uncertainty
(dBm) (dB)
300 kHz 0 + 1 + 0.47
20 MHz 0 +1 + 0.25
50 MHz 0 + 1 + 0.12
100 MHz 0 + 1 + 0.12
200 MHz 0 + 1 + 0.12
500 MHz 0 +1 + 0.12
1 GHz 0 +1 + 0.12
2 GHz 0 +1 + 0.15
3 GHz 0 +1 + 0.15
4 GHZ 0 +1 + 0.17
5 GHZ 0 +1 + 0.17
6 GHz 0 + 1 + 0.17
4 Performance Test Record
HP 8753E Performance Test Record (4 of 14)
For 30 kHz—6 GHz Analyzers
~ Hewlett-Packard Company
— Model HP 8753E Option 006 Report Number
Serial Number Date
PP 4. Test Port Output Power Range and Linearity
Test Settings Results Measured Power Level Specification | Meas. Uncert.
— (dB) Linearity (dB) (dB)
(dB)
CW Frequency = 300 kHz
— — 15 + 0.2 + 0.03
— 13 + 0.2 + 0.03
— 11 + 0.2 + 0.03
— — 9 + 0.2 + 0.02
— 7 | + 0.2 + 0.02
— 5 + 0.2 + 0.02
— — 3 . + 0.2 + 0.02
— 1 + 0.2 + 0.02
+ 1 + 0.2 + 0.03
— + 3 | + 0.2 + 0.03
+ 5 + 0.5 + 0.03
+ 7 + 0.5 + 0.03
+ 9 + 0.5 + 0.03
+ 10 + 0.5 + 0.03
CW Frequency = 3 GHz
Ш — — 15 | + 0.2 + 0.03
— 13 + 0.2 + 0.03
— 11 + 0.2 + 0.03
— 9 + 0.2 + 0.02
— 7 + 0.2 + 0.02
— 5 + 0.2 + 0.02
a — 3 + 0.2 + 0.02
— 1 + 0.2 + 0.02
_ + 1 + 0.2 + 0.03
+ 3 + 0.2 + 0.03
+ 5 + 0.5 + 0.08
Performance Test Record 5
HP 8753E Performance Test Record (5 of 14)
For 30 kHz—6 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E Option 006
Serial Number
Report Number
Date
D» 4. Test Port Output Power Range and Linearity (continued)
Test Settings Results Measured Power Level Specification | Meas. Uncert.
(dB) Linearity (dB) (dB) (dB)
+ 7 + 0.5 + 0.03
+ 9 + 0.5 + 0.03
+ 10 + 0.5 + 0.03
CW Frequency = 6 GHz
— 15 + 0.2 + 0.03
— 13 + 0.2 + 0.03
— 11 + 0.2 + 0.03
— 9 + 0.2 + 0.08
- 7 + 0.2 + 0.02
— 5 + 0.2 + 0.02
— 3 + 0.2 + 0.02
— 1 + 0.2 + 0.02
+ 1 + 0.2 + 0.02
+ 3 + 0.2 + 0.03
+ 5 + 0.5 + 0.03
+ 7 + 0.5 + 0.03
+ 9 + 0.5 + 0.03
+ 10 + 0.5 + 0.03
6 Performance Test Record
HP 8753E Performance Test Record (6 of 14)
For 30 kHz—6 GHz Analyzers
~ Hewlett-Packard Company
_— Model HP 8753E Option 006 Report Number
Serial Number Date
PP 5. Minimum R Channel Level
CW Frequency Specification Test Port Power Measurement
no (dB) Uncertainty
(dB)
— 300 kHz < —35 + 1.0
3.29 MHz < —35 + 1.0
3.31 MHz < —35 + 1.0
— 15.90 MHz < —35 + 1.0
16.10 MHz < —3b + 1.0
30.90 MHz < —35 + 1.0
_— 31.10 MHz < —35 + 1.0
1.6069 GHz < —35 + 1.0
1.6071 GHz < —35 + 1.0
— 3.000 GHZ < —35 + 2.0
4.000 GHz < —30 + 2.0
5.000 GHz < — 30 + 2.0
— 6.000 GHz < —30 + 2.0
PP 6. Test Port Input Noise Floor Level
oT Frequency Range Test Port IF Bandwidth Specification Calculated Measurement
(dBm) Value Uncertainty
300 kHz—3 GHz Port 1 3 kHz — 82 ———Ñ— N/A
300 kHz—3 GHz Port | 10 Hz — 102 ——]—]] | N/A
— 300 kHz--3 GHZ Port 2 10 Hz — 102 AAN N/A
300 kHz—3 GHz Port 2 3 kHz — 82 —_— N/A
3 GHz—6 GHz Port 2 3 kHz — TT RE N/A
3 GHz—6 GHz Port 2 10 Hz — 97 — N/A
3 GHz—6 GHz Port 1 10 Hz — 97 manana N/A
3 GHz—6 GHz Port 1 3 kHz — 77 — N/A
Performance Test Record 7
HP 8753E Performance Test Record (7 of 14)
For 30 kHz—6 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E Option 006
Serial Number
Report Number
Date
PP 7. Test Port Input Frequency Response
Frequency Range Test Port Specification Measured Value Measurement
(dB) (dB) Uncertainty
(dB)
300 kHz-—3 GHz Port 2 +1 0.47
300 kHz—3 GHz Port 1 + 1 0.47
3 GHz—6 GHz Port 1 + 2 0.17
3 GHz—6 GHz Port 2 + 2 0.17
PP 8. Test Port Crosstalk
Test Settings Specification Measured Value Measurement
(dB) (dB) Uncertainty
Crosstalk to Test Port 2
300 kHz—3 GHz < — 100 N/A
Crosstalk to Test Port 1
300 kHz—3 GHz < —100 N/A
Crosstalk to Test Port 1
à GHz—6 GHz < —90 N/À
Crosstalk to Test Port 2
3 GHz-6 GHz < —90 N/A
8 Performance Test Record
HP 8753E Performance Test Record (8 of 14)
For 30 kHz—6 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E Option 006
Serial Number
Date
Report Number
Py 9. Calibration Coeflicients
Test Description Frequency Range Spec. Measured Value Measurement
(dB) (dB) Uncertainty
| (dB)
Forward Direction
Directivity 300 kHz—1.3 GHz > 35 + 0.9
Directivity 1.3 GHz—3 GHz > 30 + 0.8
Directivity 3 GHz—6 GHz > 28 + 0.8
Forward Direction
Source Match 300 kHz—1.3 GHz > 16 + 0.2
Source Match 1.3 GHz—3 GHz > 16 + 0.2
Source Match 3 GHz—6 GHz > 14 + 0.3
Forward Direction
Trans. Tracking 300 kHz—1.3 GHz + 1.5 + 0.006
Trans. Tracking 1.83 GHz—3 GHz + 1.5 + 0.009
Trans. Tracking 3 GHz—6 GHz + 2.5 + 0.021
Forward Direction
Refl. Tracking 300 kHz—1.3 GHz + 1.5 + 0.001
Refl. Tracking 1.3 GHz—3 GHz + 1.5 + 0.005
Refl. Tracking 3 GHz—6 GHz + 2.5 + 0.020
Reverse Direction
Load Match 300 kHz-—1.3 GHz > 18 + 0.1
Load Match 1.3 GHz—3 GHz > 16 + 0.2
Load Match 3 GHz—6 GHz > 14 + 0.2
Performance Test Record 9
HP 8753E Performance Test Record (9 of 14)
For 30 kHz—6 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E Option 006
Serial Number
Report Number
Date
P» 9. Calibration Coefficients (continued)
Test Description Frequency Range Spec. Measured Value Measurement
(dB) (dB) Uncertainty
(dB)
Reverse Direction
Trans. Tracking 300 kHz—1.3 GHz + 1.5 + 0.006
Trans. Tracking 1.3 GHz—3 GHz + 1.5 + 0.009
Trans. Tracking 3 GHz—6 GHz + 2.5 + 0.021
Forward Direction
Load Match 300 kHz—1.3 GHz > 18 + 0.1
Load Match 1.3 GHz—3 GHz > 16 + 0.2
Load Match 3 GHz—6 GHz > 14 + 0.2
Reverse Direction
Directivity 300 kHz—1.3 GHz > 35 + 0.9
Directivity 1.3 GHz—à GHz > 30 + 0.8
Directivity 3 GHz—6 GHz > 25 + 0.8
Reverse Direction
Source Match 300 kHz - 1.3 GHz > 16 + 0.2
Source Match 1.3 GHz - 3 GHz > 16 + 0.2
Source Match 3 GHz - 6 GHz > 14 + 0.3
Reverse Direction
Refl. Tracking 300 kHz - 1.3 GHz + 1.5 + 0.001
Refl. Tracking 1.3 GHz - 3 GHz + 1.5 + 0.005
Refl. Tracking 3 GHz - 6 GHz + 2.5 + 0.020
10 Performance Test Record
HP 8753E Performance Test Record (10 of 14)
For 30 kHz—6 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E Option 006
Report Number
Serial Number Date
PD» 11. System Trace Noise
Frequency Ratio Measured Value Specification Measurement
(GHz) Uncertainty
3 A/R (Magnitude) < 0.006 dB rms + 0.001 dB
6 A/R (Magnitude) < 0.010 dB rms + 0.001 dB
6 A/R (Phase) < 0.070° rms + 0.01°
3 A/R (Phase) < 0.038° rms + 0.01 °
3 B/R (Magnitude) < 0.006 dB rms + 0.001 dB
6 B/R (Magnitude) < 0.010 dB rms + 0.001 dB
6 B/R (Phase) < 0.070° rms + 0.01 9
3 B/R (Phase) < 0.038° rms + 0.01 °
PP 12. Test Port Input Impedance
Test Description Test Port Return Loss Specification Measurement
(dB) (dB) Uncertainty
(dB)
300 kHz-—1.3 GHz Port 2 > 18 + 1.5
1.3 GHz—3 GHz Port 2 > 16 + 1.5
à GHz—6 GHz Port Z > 14 + 1.0
300 kHz—1.3 GHz Port 1 > 18 + 1.5
1.3 GHz—3 GHz Port 1 > 16 + 1.5
3 GHz—6 GHz Port 1 > 14 + 1.0
Performance Test Record 11
HP 8753E Performance Test Record (11 of 14)
For 30 kHz—6 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E Option 006
Serial Number
Report Number
Date
Pp» 13. Test Port Receiver Magnitude Dynamic Accuracy
G F |G — F|
Test Port 8496A Test Port Expected Dynamic Spec. Meas.
Input Power Attn. Measurement Measurement Accuracy (dB) Uncer.
(dB) (dBm) (corrected) (Calculated) (dB)
(dBm)
Test Port 2
— 10 0 < 0.033 + 0.008
— 20 (Ref) 10 Reference Reference Reference Reference Reference
— 30 20 < 0.031 + 0.008
— 40 30 < 0.042 + 0.008
— 50 40 < 0.057 + 0.008
— 60 50 < 0.098 + 0.017
— 70 60 < 0.247 + 0.017
— 80 70 < 0.725 + 0.017
— 90 80 < 2.097 + 0.017
— 100 90 < 5.399 + 0.027
Test Port 1
— 10 0 < 0.033 + 0.008
— 20 (Ref) 10 Reference Reference Reference Reference Reference
— 30 20 < 0.031 + 0.008
— 40 30 < 0.042 + 0.008
— 50 40 < 0.057 + 0.008
— 60 50 < 0.098 + 0.017
— 70 60 < 0.247 + 0.017
— 80 70 < 0.725 + 0.017
— 90 80 < 2.097 + 0.017
— 100 90 < 5.399 + 0.027
12 Performance Test Record
HP 8753E Performance Test Record (12 of 14)
For 30 kHz—6 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E Option 006
Serial Number
Report Number
Date
bb 14. Test Port Receiver Magnitude Compression
CW Frequency Test Port Measured Value Specification Measurement
(dB) (dB) Uncertainty
50 MHz Port 2 < 0.45 N/A
1 GHZ Port 2 < 0.45 N/A
2 GHz Port 2 < 0.45 N/A
3 GHz Port 2 < 0.45 N/A
4 GHz Port 2 < 0.80 N/A
5 GHZ Port 2 < 0.80 N/A
6 GHZ Port 2 < 0.80 N/A
50 MHz Port 1 < 0.45 N/A
1 GHZ Port 1 < 0.45 N/A
2 GHz Port 1 < 0.45 N/A
3 GHZ Port 1 < 0.45 N/A
4 GHz Port 1 < 0.80 N/A
5 GHz Port 1 < 0.80 N/A
6 GHz Port 1 < 0.80 N/A
Performance Test Record
13
HP 8753E Performance Test Record (13 of 14)
For 30 kHz—6 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E Option 006
Serial Number
Report Number
Date
PP 15. Test Port Receiver Phase Compression
CW Frequency Test Port Measured Value Specification Measurement
(degrees) (degrees) Uncertainty
50 MHz Port 2 < 6° N/A
1 GHz Port 2 < 6° N/A
2 GHz Port 2 < 6° N/A
3 GHz Port 2 < 6° N/A
4 GHz Port 2 < 7.5" N/A
5 GHz Port 2 < 7.5% N/A
6 GHz Port 2 < 7.59 N/A
50 MHz Port 1 < 6° N/A
1 GHz Port 1 < 6° N/A
2 GHZ Port 1 < 6° N/A
3 GHz Port 1 < 6° N/A
4 GHz Port 1 < 7.59 N/A
5 GHz Port 1 < 7.5% N/A
6 GHZ Port 1 < 7.5% N/A
14 Performance Test Record
HP 8753E Performance Test Record (14 of 14)
For 30 kHz—6 GHz Analyzers
Hewlett-Packard Company
Model HP 8753E Option 006 Report Number
Serial Number Date
»» 17. Output/Input Test Port Harmonics (Option 002 only)
Test Description Specification Measurement Value Measurement Uncertainty
(dBc) (dBc) (dB)
Test Port Output Harmonics
2nd < 25 + 1.5
¿rd < 25 + 1.5
Port 1 Input Harmonics
2nd < 15 + 1.5
ard < 30 + 1.5
Port 2 Input Harmonics
2nd < 15 + 1.5
ard < 30 + 1.5
Performance Test Record 15
Index
A
accuracy and range of frequency, 1-11
accuracy of power test, 1-15
analyzer verification, 1-1
C
calibration certificate, 1-2
certificate of calibration, 1-2
certification of kit, 1-4
E
equipment
automated system verification, 1-5
external source mode frequency range,
1-13
minimum R channel level, 1-21
test port frequency range and accuracy
test, 1-11
test port input noise floor level, 1-26
test port output power accuracy, 1-15
test port output power range and linearity,
1-18
external source mode frequency range, 1-13
F
floor level test, 1-26
frequency range and accuracy test, 1-11
frequency range for external source mode,
1-13
H
how to
performance test the analyzer, 1-2
test external source mode frequency range,
1-13
test frequency range and accuracy, 1-11
test minimum R channel level, 1-21
test port input noise floor level, 1-26
test port output frequency range and
accuracy, 1-11
test port output power accuracy, 1-15
test port output power range and linearity,
1-18
verify analyzer operation, 1-2
verify an analyzer system automatically,
1-5
I
input noise floor level test, 1-26
K
kit re-certification, 1-4
L
linearity and range of power test, 1-18
M
minimum R channel level, 1-21
N
noise floor level test, 1-26
O
operation verification, 1-1
P
performance test record types, 1-3
performance tests
1. Test Port Output Frequency Range and
Accuracy, 1-11
2. External Source Mode Frequency Range,
1-13 |
3. Test Port Output Power Accuracy, 1-15
4, Test Port Output Power Range and
Linearity, 1-18
5. Minimum R Channel Level, 1-21
6. Test Port Input Noise Floor Level, 1-26
chapter, 1-1
description of, 1-1
port input noise floor level test, 1-26
port output power accuracy test, 1-15
power accuracy test, 1-15
power range and linearity test, 1-18
procedures
external source mode frequency range,
1-13
minimum R channel level, 1-21
Test Port Input Noise Floor Level, 1-26
test port output frequency range and
accuracy, 1-11
test port output power accuracy, 1-15
Index-1
test port output power range and linearity,
1-18
verify an analyzer system (automated),
1-5
R
range and accuracy of frequency, 1-11
R channel level, 1-21
S
setup
external source mode frequency range,
1-13
minimum R channel level, 1-21
mismatch device verification, 1-9
test port frequency range and accuracy
test, 1-12
test port input noise floor level, 1-27
test port output power accuracy, 1-16
test port output power range and linearity,
1-19
transmission calibration, 1-7
source mode frequency range, 1-13
specifications
external source mode frequency range,
1-13
minimum R channel level, 1-21
test port input noise floor level, 1-26
test port output frequency range and
accuracy, 1-11
index-2
test port output power accuracy, 1-15
test port output power range and linearity,
1-18
system verification
description of, 1-1
system verification (automated), 1-5
system verification cycle, 1-4
T
test port input noise floor level, 1-26
test port output frequency range and accuracy
test, 1-11
test port output power accuracy, 1-15
test port output power range and linearity,
1-18
test record types, 1-3
tests
1. Test Port Output Frequency Range and
Accuracy, 1-11
2. External Source Mode Frequency Range,
1-13
chapter, 1-1
minimum R channel level, 1-21
Test Port Input Noise Floor Level:, 1-26
Test Port Output Power Accuracy, 1-15
Test Port Output Power Range and Linearity,
1-18
V
verification cycle, kit re-certification, 1-4
">
/
Herunterladen
Nur eine freundliche Erinnerung. Sie können das Dokument direkt hier ansehen. Aber was am wichtigsten ist, unsere KI hat es bereits gelesen. Sie kann komplexe Dinge einfach erklären, Ihre Fragen in jeder Sprache beantworten und Ihnen helfen, auch in den längsten oder kompliziertesten Dokumenten schnell zu navigieren.
Werbung