Keysight Technologies
34401A Digital Multimeter
Data Sheet
Product Discontinuance Notice
The 34401A DMM and all associated options will be discontinued December 1, 2016.
The last date this product can be ordered is
November 30, 2016.
– For new product purchases, Keysight recommends the 34461A, 6½ digit Truevolt DMM.
– For more information, as well as to access to detailed migration guides, please visit www.keysight.com/find/nextgenDMMs
– To contact a product selection expert, visit www.keysight.com/find/contactus
02 | Keysight | 34401A Digital Multimeter - Data Sheet
Measurement capability
– 6½ digit resolution
– 10 measurement functions: DC/AC voltage, DC/AC current,
2- and 4-wire resistance, diode, continuity, frequency, period
– Basic accuracy: 0.0035% DC, 0.06% AC
– 1000 V max voltage input, 3 A max current input
Use it for systems testing
For systems use, the 34401A gives you faster bus throughput than any other DMM in its class. The 34401A can send up to
1,000 readings/s directly across GPIB in user-friendly ASCII format.
System Capability
– 1000 readings/second
– 512 reading memory www.keysight.com/find/dmm
Superior performance
The Keysight Technologies, Inc. 34401A multimeter gives you the performance you need for fast, accurate bench and system testing. The 34401A provides a combination of resolution, accuracy and speed that rivals DMMs costing many times more. 6½ digits of resolution, 0.0015% basic 24-hr dcV accuracy and 1,000 readings/s direct to GPIB assure you of results that are accurate, fast, and repeatable.
You also get both GPIB and RS-232 interfaces as standard features. Voltmeter Complete and External Trigger signals are provided so you can synchronize to other instruments in your test system. In addition, a TTL output indicates Pass/Fail results when limit testing is used.
To ensure both forward and backward compatibility, the 34401A includes three command languages (SCPI, Keysight 3478A and
Fluke8840A/42A), so you don’t have to rewrite your existing test software. An optional rack mount kit is available.
Easy to use
Commonly accessed attributes, such as functions, ranges, and resolution are selected with a single button press.
Use it on your benchtop
The 34401A was designed with your bench needs in mind. Functions commonly associated with bench operation, like continuity and diode test, are built in. A Null feature allows you to remove lead resistance and other fixed offsets in your measurements.
Other capabilities like min/max/avg readouts and direct dB and dBm measurements make checkout with the 34401A faster and easier.
The 34401A gives you the ability to store up to 512 readings in internal memory. For troubleshooting, a reading hold feature lets you concentrate on placing your test leads without having to constantly glance at the display.
Advanced features are available using menu functions that let you optimize the 34401A for your applications.
The included Keysight IntuiLink software allows you to put your captured data to work easily, using PC applications such as
Microsoft Excel or Word to analyze, interpret, display, print, and document the data you get from the 34401A. You can specify the meter setup and take a single reading or log data to the Excel spreadsheet in specified time intervals. Programmers can use
ActiveX components to control the DMM using SCPI commands.
To find out more about IntuiLink, visit www.keysight.com/find/intuilink
3-year warranty
With your 34401A, you get full documentation, a high-quality test lead set, calibration certificate with test data, and a 3-year warranty, all for one low price.
03 | Keysight | 34401A Digital Multimeter - Data Sheet
Accuracy Specifications ± (% of reading + % of range)
1
Function
DC voltage
True rms AC voltage 4
Resistance
DC current
True rms
AC current
Frequency or period
Continuity
8
Diode test 9
7
4
Range 3
100.0000 mV
1.000000 V
10.00000 V
100.0000 V
1000.000 V
100.0000 mV
1.000000 V to
750.000 V
100.0000 Ω
1.000000 kΩ
10.00000 kΩ
100.0000 kΩ
1.000000 MΩ
10.00000 MΩ
100.0000 MΩ
10.00000 mA
100.0000 mA
1.000000 A
3.00000 A
1.000000 A
3.00000 A
100 mV to 750 V
1000.0 Ω
1.0000 V
Frequency, etc.
3 – 5 Hz
5 – 10 Hz
10 Hz – 20 kHz
20 – 50 kHz
50 – 100 kHz
100 – 300 kHz 6
3 – 5 Hz
5 – 10 Hz
10 Hz – 20 kHz
20 – 50 kHz
50 – 100 kHz 5
100 – 300 kHz 6
1 mA current source
1 mA
100 µA
10 µA
5.0 µA
500 nA
500 nA II 10 MΩ
< 0.1 V burden voltage
< 0.6 V
< 1.0 V
< 2.0 V
3 – 5 Hz
5 – 10 Hz
10 Hz – 5 kHz
3 – 5 Hz
5 – 10 Hz
10 Hz – 5 kHz
3 – 5 Hz
5 – 10 Hz
10 – 40 Hz
40 Hz – 300 kHz
1 mA test current
1 mA test current
0.005 + 0.010
0.010 + 0.010
0.050 + 0.010
0.100 + 0.010
1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.10 + 0.06
0.35 + 0.06
0.15 + 0.06
0.10
0.05
0.03
0.006
0.002 + 0.030
0.002 + 0.010
24 hour 2
23 ±1 °C
0.0030 + 0.0030
0.0020 + 0.0006
0.0015 + 0.0004
0.0020 + 0.0006
0.0020 + 0.0006
1.00 + 0.03
0.35 + 0.03
0.04 + 0.03
0.10 + 0.05
0.55 + 0.08
4.00 + 0.50
1.00 + 0.02
0.35 + 0.02
0.04 + 0.02
0.10 + 0.04
0.55 + 0.08
4.00 + 0.50
0.0030 + 0.0030
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0005
0.002 + 0.001
0.015 + 0.001
0.300 + 0.010
0.030 + 0.020
0.030 + 0.005
0.080 + 0.010
0.120 + 0.020
1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.10 + 0.06
0.35 + 0.06
0.15 + 0.06
0.10
0.05
0.03
0.01
0.008 + 0.030
0.008 + 0.020
90 day
23 ±5 °C
0.0040 + 0.0035
0.0030 + 0.0007
0.0020 + 0.0005
0.0035 + 0.0006
0.0035 + 0.0010
1.00 + 0.04
0.35 + 0.04
0.05 + 0.04
0.11 + 0.05
0.60 + 0.08
4.00 + 0.50
1.00 + 0.03
0.35 + 0.03
0.05 + 0.03
0.11 + 0.05
0.60 + 0.08
4.00 + 0.50
0.008 + 0.004
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.020 + 0.001
0.800 + 0.001
1 year
23 ±5 °C
0.0050 + 0.0035
0.0040 + 0.0007
0.0035 + 0.0005
0.0045 + 0.0006
0.0045 + 0.0010
1.00 + 0.04
0.35 + 0.04
0.06 + 0.04
0.12 + 0.05
0.60 + 0.08
4.00 + 0.50
1.00 + 0.03
0.35 + 0.03
0.06 + 0.03
0.12 + 0.04
0.60 + 0.08
4.00 + 0.50
0.010 + 0.004
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.040 + 0.001
0.800 + 0.001
0.050 + 0.020
0.050 + 0.005
0.100 + 0.010
0.120 + 0.020
1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.10 + 0.06
0.35 + 0.06
0.15 + 0.06
0.10
0.05
0.03
0.01
0.010 + 0.030
0.010 + 0.020
Temperature coefficient
0 – 18 °C
28 – 55 °C
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.100 + 0.004
0.035 + 0.004
0.005 + 0.004
0.011 + 0.005
0.060 + 0.008
0.20 + 0.02
0.100 + 0.003
0.035 + 0.003
0.005 + 0.003
0.011 + 0.005
0.060 + 0.008
0.20 + 0.02
0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1500 + 0.0002
0.0020 + 0.0020
0.0020 + 0.0005
0.0050 + 0.0010
0.005 + 0.0020
0.100 + 0.006
0.035 + 0.006
0.015 + 0.006
0.100 + 0.006
0.035 + 0.006
0.015 + 0.006
0.005
0.005
0.001
0.001
0.001 + 0.002
0.001 + 0.002
1. Specifications are for 1 hr warm-up and 6½ digits, slow ac filter.
2. Relative to calibration standards.
3. 20% over range on all ranges except 1000 Vdc and 750 Vac ranges.
4. For sinewave input > 5% of range. For inputs from 1% to 5% of range and < 50 kHz, add 0.1% of range additional error.
5. 750 V range limited to 100 kHz or 8 x 107 Volt-Hz.
6. Typically 30% of reading error at 1 MHz.
7. Specifications are for 4-wire ohms function or 2-wire ohms using Math Null. Without Math Null, add 0.2 Ω additional error in 2-wire ohms function.
8. Input > 100 mV. For 10 to 100 mV inputs multiply % of reading error x 10.
9. Accuracy specifications are for the voltage measured at the input terminals only. 1 mA test current is typical. Variation in the current source will create some variation in the voltage drop across a diode junction.
254.4 mm 374.0 mm
212.6 mm 348.3 mm 254.4 mm
212.6 mm
374.0 mm
348.3 mm
04 | Keysight | 34401A Digital Multimeter - Data Sheet
Measurement Characteristics
DC voltage
Measurement method
A-D linearity
Input resistance
10 MΩ or 0.1V, 1 V, 10 V ranges
100 V, 1000 V ranges
Input bias current
Input protection dcV:dcV ratio accuracy
Continuously integrating multi-slope III A-D converter
0.0002% of reading + 0.0001% of range
Selectable > 10,000 MΩ
10 MΩ ±1%
< 30 pA at 25°C
1000 V all ranges
V input
accuracy + V relevance
accuracy
True RMS AC voltage
Measurement method
Crest factor
Additional crest factor errors (non-sinewave)
Input impedance
Input protection
Resistance
Measurement method
Maximum lead resistance
(4-wire)
Input protection
DC current
Shunt resistance
Input protection
Frequency and period
Measurement method
Voltage ranges
Gate time
Continuity/diode
Response time
Continuity threshold
Reciprocal counting technique
Same as ac voltage function
1 s, 100 ms, or 10 ms
300 samples/s with audible tone
Selectable from 1 Ω to 1000 Ω
Measurement noise rejection 60 (50) Hz dc CMRR ac CMRR
140 dB
70 dB
1
AC-coupled true rms-measures the ac component of the input with up to 400 Vdc of bias on any range
Maximum of 5:1 at full scale
Crest factor 1-2: 0.05% of reading
Crest factor 2-3: 0.15% of reading
Crest factor 3-4: 0.30% of reading
Crest factor 4-5: 0.40% of reading
1 MΩ ± 2% in parallel with 100 pF
750 Vrms all ranges
Selectable 4-wire or 2-wire Ohms
Current source referenced to LO input
10% of range per lead for 100 Ω, 1 kΩ ranges
1 kΩ per lead on all other ranges
1000 V all ranges
5 Ω for 10 mA, 100 mA
0.1 Ω for 1 A, 3 A
Externally accessible 3 A 250 V fuse
Internal 7 A 250 V fuse
Integration time and normal mode rejection 2
100 plc/1.67 s (2 s) 60 dB 3
0 plc/167 ms (200 ms)
1 plc/16.7 ms (20 ms)
< 1 plc/3 ms or 800 µs)
60 dB
60 dB
0 dB
3
Operating characteristics 4
Function dcV, dcl, and resistance acV, acl
Frequency or period
6½
6½
6½
5½
4½
5½
4½
6½
6½
Digits
6½
6½
5½
Reading/s
0.6 (0.5)
6 (5)
60 (50)
300
1000
0.15 slow (3 Hz)
1 medium (20 Hz)
10 fast (200 Hz) 5
50
1
9.8
80
1. For 1 kΩ unbalanced in LO lead, ± 500 V peak maximum.
2. For power line frequency ± 0.1%.
3. For power line frequency ± 1% use 40 dB or ± 3% use 30 dB.
4. Reading speeds for 60 Hz and (50 Hz) operation.
5. Maximum useful limit with default settling delays defeated.
6. Speeds are for 4½ digits, delay 0, auto-zero and display OFF.
True RMS AC current
Measurement method
Shunt resistance
Input protection
Directly coupled to the fuse and shunt. ac coupled true rms measurement (measures the ac component only).
0.1 Ω for 1 A and 3 A ranges
Externally accessible 3 A 250 V fuse
Internal 7 A 250 V fuse
05 | Keysight | 34401A Digital Multimeter - Data Sheet
Measurement Characteristics
System speeds
Configuration rates
Autorange rate (dc Volts)
26/s to 50/s
> 30/s
ASCII readings to RS-232 55/s
ASCII readings to GPIB 1000/s
Maximum internal trig rate 1000/s
Max. ext trig. rate to mem 1000/s
continued
Triggering and memory
Reading HOLD sensitivity 10%, 1%, 0.1%, or 0.01% of range
Samples/trigger
Trigger delay
1 to 50,000
0 to 3600 s: 10 µs step size
External trigger delay
External trigger jitter
Memory
< 1 ms
< 500 µs
512 readings
Math functions
NULL, min/max/average, dBm, dB, limit test (with TTL output)
Standard programming languages
SCPI (IEEE-488.2), Keysight 3478A, Fluke 8840A/42A
Accessories included
Test lead kit with probe, alligator and grabber attachments
Operating manual, service manual, test report and power cord
General specifications
Power supply
Power line frequency
Power consumption
Operating environment
Storage temperature
Weight
Safety
RFI and ESD
Vibration and shock
Warranty
100 V/120 V/220 V/240 V ±10%
45 to 66 Hz and 360 to 440 Hz, automatically sensed at power-on
25 VA peak (10 W average)
Full accuracy for 0 to 55 °C, full accuracy to
80% R.H. at 40 °C
–40 to 70 °C
3.6 kg (8.0 lbs)
Designed to CSA, UL-1244, IEC-348
MIL-461C, FTZ 1046, FCC
MIL-T-28800E, Type III, Class 5 (sine only)
1 year
Ordering Information
Keysight 34401A multimeter accessories included: Test lead kit with probe, alligator, and grabber attachments, calibration certificate, test report, and power cord.Also includes CD with:
IntuiLink software, IVI and VXI PnP drivers, Quick start tutorial, user’s guide, command quick reference, service guide, and data sheet.
Options
34401A-A6J ANSI Z540 compliant calibration
Accessories
Probes/Leads/Clip Accessories
11059A Kelvin probe set
11060A
11062A
Surface mount device (SMD) test probes
Kelvin clip set
34133A
34134A
34136A
34138A
Precision electronic test leads
DC coupled current probe
High voltage probe
Test lead set
34171B
34172B
34330A
E2308A
Y1133A
Input terminal connector (sold in pairs)
Input calibration short (sold in pairs)
30 A current shunt
5 k thermistor probe
Low-thermal external digital multimeter scanning kit
Rackmount kits
34190A Rackmount kit Designed for use with only one instrument, mounted on either the left or the right side of the rack.
34191A 2U Dual flange kit
Secures the instrument to the front of the rack.
This kit can be used with the 34194A dual lock link kit to mount two half-width, 2U height instruments side-by side.
34194A Dual lock link kit
Recommended for side-by-side combinations and includes links for instruments of different depths. This kit can be used with the 34191A
2U dual flange kit to mount two half-width, 2U height instruments side-by-side.
Other accessories
34131A
34161A
34398A
E5810B
Hard transit case
Accessory pouch
RS-232 cable, 9 pin (f) to 9 pin (f)
LAN/GPIB gateway
06 | Keysight | 34401A Digital Multimeter - Data Sheet myKeysight www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
www.axiestandard.org
AdvancedTCA ® Extensions for Instrumentation and Test (AXIe) is an open standard that extends the AdvancedTCA for general purpose and semiconductor test. Keysight is a founding member of the AXIe consortium.
ATCA ® , AdvancedTCA ® , and the ATCA logo are registered US trademarks of the PCI Industrial Computer Manufacturers Group. www.lxistandard.org
LAN eXtensions for Instruments puts the power of Ethernet and the
Web inside your test systems. Keysight is a founding member of the LXI consortium.
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2008
Quality Management System
Keysight Channel Partners www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience.
www.keysight.com/find/34401A
For more information on Keysight
Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
55 11 3351 7010
001 800 254 2440
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 11 2626
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 6375 8100
Europe & Middle East
Austria
Belgium
Finland
France
Germany
Ireland
Israel
Italy
Luxembourg
Netherlands
Russia
Spain
Sweden
Switzerland
United Kingdom
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
800 000154
0200 882255
0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
0800 0260637
For other unlisted countries: www.keysight.com/find/contactus
(BP-04-23-15)
This information is subject to change without notice.
© Keysight Technologies, 2013 - 2015
Published in USA, May 21, 2015
5968-0162EN www.keysight.com