Sibelmed DATOSPIR TOUCH DIAGNOSTIC - F User manual
Reklama
Reklama
DATOSPIR TOUCH
SPIROMETER
USER’S MANUAL
511-B00-MU2 • REV. 1.02 • 2011/09
DATOSPIR TOUCH User’s Manual 2
1
SIBEL S.A., Rosselló 500 bajos, 08026 Barcelona - Spain
National Sales: Tel. 93 436 00 08 e-mail: [email protected]
International Sales: Tel. +34 93 436 00 07 e-mail: [email protected]
Technical service: Tel. +34 93 433 54 50 e-mail: [email protected]
Fax: +34 93 436 16 11, Website:
www.sibelmed.com
SIBEL, S.A. belongs to SIBELGROUP
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 3
1
1. SAFETY…………………………………………………………… 5
2. INSTRUCTIONS FOR USE & INSTALLATION ..….…… 10
2.1. INTRODUCTION …………………………………………………………………… 10
2.2. MODELS AND OPERATING MODES ……………………………………… 10
2.3. LAYOUT OF CONTROLS AND CONNECTORS ……………………… 17
2.4. INSTALLATION AND START-UP …………………………………………… 18
3. SPIROMETER CONFIGURATION ………………………… 23
3.1. SETUP ………………………………………………………………………………… 23
3.2. CUSTOMIZATION ………………………………………………………………… 23
3.3. INTERNAL DATABASE…………………………………………………………… 29
3.4. MAINTENANCE PROGRAM …………………………………………………… 30
3.5. QUALITY CONTROL: CALIBRATION CHECK …………………………31
3.6. CALIBRATION ROCEDURE …………………………………………………… 34
3.7. CALIBRATION RECORD …………………………………………………………35
4. SPIROMETRIC TESTS PROCEDURES …………………… 36
4.1. FORCED VITAL CAPACITY “FVC” TEST ……………………………… 36
4.2. QUALITY OF FVC TESTS ……………………………………………………… 45
4.3. POST BRONCHIALDILATOR TEST………………………………………… 48
4.4. SLOW VITAL CAPACITY “VC” TEST …………………………………… 50
4.5. MAXIMUM VOLUNTARY VENTILATION “MVV” TEST …………… 51
5. COMMUNICATIONS SYSTEM ……………………………… 52
5.1. TRANSFERRING TESTS TO THE PC …………………………………… 52
5.2. TRANSFERRING EQUIPMENT CHECK DATA …………………………53
5.3. ADDING MODULE, OPTIONS AND/OR TRANSDUCER …………54
5.4. UPDATING FIRMWARE …………………………………………………………55
5.5. EXPORTING TESTS TO OTHER SYSTEMS ……………………………55
6. TECHNICAL SPECIFICATIONS ……………………..…..…57
6.1. GENERAL SPECIFICATION ……………………………………………………57
6.2. SYMBOLS ………………………………………………………………………………57
6.3. TESTS, FUNCTIONS AND PARAMETERS ………………………………59
6.4. PREDICTED SETS …………………………………………………………………60
6.5. FVC INTERPRETATION (DIAGNOSIS) ………………………………… 62
6.6. TRANSDUCERS …………………………………………………………………… 62
6.7. MANUFACTURER’S LIABILITY ………………………………………………63
7. CLEANING AND MAINTENANCE ……………………..……64
7.1. CLEANING / DISINFECTION ……………………………………………… 64
7.2. PREVENTIVE MAINTENANCE ……………………………………………… 66
7.3. CORRECTIVE MAINTENANCE ……………………………………………… 67
Annex 1. ELECTROMAGNETIC COMPATIBILITY …..……. 68
Annex 2. COMPLIANCE WITH THE DATA
PROTECTION ACT. DIRECTIVE 95/46/EC …………..…… 72
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 4
1
The DATOSPIR TOUCH Spirometer has been designed by the
R+D+I Department of SIBEL S.A., with the collaboration of the Pneumology Service of Hospital de la Santa Creu i Sant
Pau de Barcelona, in line with the standardization criteria of
International Institutions: ATS/ERS TASK FORCE 2005
(American Thoracic Society/ European Respiratory Society) and
National Institutions: SEPAR (Spanish Pneumology and Thoracic
Surgery Society).
COMPLIANT PRODUCT
93/42/EEC Medical Device Directive.
Class II a
Revised Approved
Date: 2011/09
Technical Director Sales Director
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 5
1
1. SAFETY
SPECIAL PRECAUTIONS
The DATOSPIR TOUCH spirometer has been designed for use with the safety in mind. All operating instructions must be read before using it. Failure to do so could cause injury to the user or patient and damage to the equipment and/or accessories.
INTENDED USE
a) Measurement of lung flows and volumes for the diagnostic and control of respiratory diseases (Asthma, COPD, etc.). b) Measurement of peripheral blood oxygen saturation and cardiac pulse for the respiratory diagnostic. c) Measurement of inspiratory and expiratory maxima pressures for the respiratory diagnostic.
Use in a health center or similar and indoor use (not for outdoor use). Not intended for home use or for use in moving transport vehicles.
INDICATIONS FOR USE
The spirometer should NOT be used under high ambient noise levels to ensure that the patient may hear the acoustic signal of the device.
The equipment should be placed in a safe position to avoid falling that may result in equipment damage or harm to the patient and/or user.
The spirometer is NOT designed for use under other conditions or using other power sources not indicated in this manual. Use only accessories specified in this manual.
USER PROFILE
The spirometer is intended to be used by or under the direction of a medical professional. Specific training on the Spirometry technique is recommended.
Bronchoconstriction test must be supervised by a qualified technician in the art.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 6
1
Before using the spirometer on patients, you should be familiar with the operation of equipment. All information necessary for its operation is available in this Manual.
For additional training on the technique or on the product, contact
SIBEL S.A. or your dealer.
EFFECTS ON PATIENTS USING THE SPIROMETER
The spirometry tests require patient cooperation.
Complete forced expiration is required to obtain meaningful patient
FVC values. The clinician administering the test must assess the patient’s capacity to perform the spirometry test. Special attention must be paid to children, the elderly and the disabled.
LIMITATIONS FOR USE. CONTRAINDICATIONS
An analysis of the results of spirometry tests is not enough to give a correct diagnosis of the patient’s clinical condition. The patient’s records and any tests that the clinician believes necessary must therefore also be considered. A doctor must interpret all data to determine the course of treatment required.
The patient’s symptoms and capacity to perform a spirometry test must be taken into account by medical staff before any spirometric testing is undertaken. Acceptability of a test is the responsibility of the medical professionals.
The spirometer should not be used when it is likely that the validity of the results could be compromised by external factors.
Take care NOT to place the equipment where it could be splashed by water or other liquids or cover it with objects that prevent air from circulating around it while it is running.
The device should NOT be used stacked or adjacent to other equipment.
All accessories and spare parts must be originals and they must be requested from the manufacturer or authorized dealer in order to ensure the safety of the patient and the correct working order of the spirometer. Failure to do so may result in an increase of emissions or in a decrease of the immunity of the equipment.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 7
1
The equipment must be stored and used within the temperature, pressure and humidity ranges specified in the section
6.1.
WATER INGRESS PROTECTION LEVEL
IPX2. Equipment Protected against falling water equivalent to 3-
5mm rainfall per minute during 10 minutes. Unit placed tilted 15 degrees, in each direction, from normal operating position. In compliance with EN-ISO 9919:2009.
ELECTRICAL RISKS
To avoid the risk of electric shock, this equipment must be connected to a power supply with protective earth only.
DO NOT tamper with the integrity of the system's electric earth connection. Protection against electrical discharge is provided by the connection of the chassis to an electrical earth connection. The earth connection is only effective when the three-wire power cable supplied with the equipment is connected to a suitably earthed electrical socket.
DO NOT use multiple mains sockets to NOT degrade the electrical safety.
DO NOT disassemble the equipment or accessories casing. The device must only be serviced and repaired by skilled personnel. The contact with voltage inside the device may cause serious injury.
DO NOT connect a line phone to the MIP-MEP connector.
DO NOT use damaged accessories. DO NOT use the equipment if the power cable is in poor condition or cracked.
ELECTRICAL DISCHARGE
To ensure vital safety features under the EN 60601-1 standard, only equipment compliant with the electrical safety standards in force may be connected to this device. To connect DATOSPIR TOUCH to a non-medical device as a printer or PC, they must be compliant with EN60950 and the installation must provide an additional safeguard. This safeguard could be an additional ground conductor connected to the metallic enclosure of the non-medical device
(contact to Technical Support to see if it is practicable) or a USB separation device provided by the manufacturer SIBEL S.A.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 8
1
NEVER immerse any part of the equipment in liquid.
THIS COULD
CAUSE AN ELECTRIC SHOCK. Consult the section
7.1.
RISKS OF EXPLOSION
NOT suitable for use in the presence of volatile anesthetics, flammable gases or in oxygen-rich environments. THIS MAY CAUSE
AN EXPLOSION.
RISKS OF CONTAMINATION
To avoid the risk of contamination or cross infection, the Turbine and Fleisch transducers must be disinfected before use (consult the section
7.1
) or used with a protective barrier filter, compatible with the equipment.
Reusable mouthpieces must also be disinfected. Disposable transducers and disposable mouthpieces must NOT be reused.
DO NOT use mouthpieces or other supplies from manufacturers that have not tested their biocompatibility, since it could endanger patient health.
RISKS OF INTERFERENCE
This is an electronic medical device; therefore requires special precautions regarding electromagnetic compatibility (EMC): it must be installed and put in service according to the information attached in
Annex 1. ELECTROMAGNETIC COMPATIBILITY.
As this is an electronic product, high frequency emissions may interfere with its correct use. Thus, keep the spirometer away from products that may generate interference (radios, cell phones, etc).
All accessories, transducers, cables and spare parts must be original and must be requested from the manufacturer or dealer, to ensure patient safety and ensuring the proper operation of the spirometer. Failure to do so may increase the emissions or decrease the immunity of the equipment.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 9
1
DISPOSAL OF ELECTRICAL AND ELECTRONIC DEVICES BY
DOMESTIC USERS IN THE EUROPEAN UNION
Never dispose of the DATOSPIR TOUCH, its accessories and its batteries in the household trash. It must be disposed of properly and may need to be recycled in accordance with the statutory requirements in your country.
The device contains lead for the electric soldering and PBB in the plastic cover.
It uses a lithium battery and could use an optional NiMh battery.
Information on proper disposal is available from your dealer or from Technical Support at SIBEL S.A.
511-B00-MU2 • REV. 1 .02
1
2. INSTRUCTIONS FOR USE AND INSTALLATION
2.1 INTRODUCTION
The DATOSPIR TOUCH spirometer is a compact device based on different ty pes of transducers: Fleisch, Turbine or Lilly
disposable, a wide high-resolution color touch screen and an internal thermal printer. Has an internal database to store the performed tests and also allows connection to an external printer via USB or Bluetooth. In addition, it may incorporate a Weather
Station to measure pressure and humidity (it includes a
temperature sensor); a MIP-MEP module to measure Maximal
Inspiratory and Expiratory Pressures, an electronic Pulse oximetry module, exclusively for taking Oxygen Saturation and Pulse Rate samples (SpO
2
) and an Ethernet module.
Furthermore, the spirometer may be connected in real or deferred time to a PC via USB, Bluetooth or
Ethernet
, using the W20s
Spirometry software to perform tests, download tests results from the device to permanent records, print hard copies of selected data or transfer test to telemedicine platforms or clinical information systems. Thus, it allows carrying out interactive real-time testing on your PC and transfer status device data
2.2 MODELS AND OPERATING MODES
2.2.1. MODELS
The DATOSPIR TOUCH series is available in 6 different models:
• DATOSPIR TOUCH EASY - T • DATOSPIR TOUCH DIAGNOSTIC - T
• DATOSPIR TOUCH EASY - F • DATOSPIR TOUCH DIAGNOSTIC - F
• DATOSPIR TOUCH EASY - D • DATOSPIR TOUCH DIAGNOSTIC - D
This manual is intended for all models and options of the
DATOSPIR TOUCH spirometer. Therefore, only specific options and functions of the model available will be applicable in each case.
511-B00-MU2 • REV. 1 .02
1
The following table shows the standard features and optional functions for each model.
The spirometer has three possible different transducers:
• Turbine Transducer
• Fleisch Transducer
• Disposable Transducer (Lilly type)
And two software configuration choices:
• Easy – for occupational and primary care
• Diagnostic – for lung function laboratories (occupational and primary care also included).
Easy Diagnostic
TRANSDUCERS
D T F D T F
Disposable
Turbine
Fleisch
SETTINGS
Occupational Medicine Mode
Primary care Mode
Diagnostic mode
1.000 test internal database with graphics
3.000 test internal database with graphics
Bronchoconstriction module
W20s Spirometry software
CONNECTIVITY
USB to extern printer connection
USB to PC
Ethernet module
Bluetooth module
511-B00-MU2 • REV. 1 .02
1
2.2.2. OPERATING MODES
There are 3 operating modes available. Choose the settings that best fit your needs:
• Diagnostic Mode (DG): Orientated to
Lung Function
Laboratories
. This is the most complete mode. Practically all the functionalities of the device are available and it allows you to use
whatever o perating mode.
• Primary care Mode (PC): Orientated to Primary Care. It includes quality alerts to assist you in producing high quality spirometry tests . Direct access to FVC, Bronchodilatation, VC and MVV tests.
• Occupational Medicine
Mode (OC):
Orientated to Prevention
Centers and Mutual. Ideal for "screening", it is easy to use and allows you to perform FVC and Bronchodilatation tests.
511-B00-MU2 • REV. 1 .02
FUNCTIONALITIES
FVC
VC
MVV
Bronchodilatation (Post)
Bronchoconstriction
Alternated V-T & F-V plots
Simultaneous V-T & F-V plots
Superposed curves
Acoustic signal for begin and end of maneuver
Time-based progress bar (adult incentive)
Volume-based progress bar (adult incentive)
Parameter and plot selection (customization)
Calibration program
Cal-check
Large graphs on reports
Print/save 3 maneuvers
Print 3 PRE maneuvers (data and graphs)
Time-audit module
Miller Interpretation
Snider, Kory&Lyons, NLHEP (Ferguson) Interpretation
ATS/ERS (McKay) Interpretation
Quality of the FVC test (NLHEP: QC prompts, QC grades)
Help screen in all menus (text)
Pediatric/adult incentives (in device)
Device auto-check program
Import external patient data
Export data to Health Information Systems
A model can be upgraded at any time by adding the corresponding parts. To do so, please contact the SIBEL S.A. Sales Department or your dealer.
511-B00-MU2 • REV. 1 .02
1
PC OC DIAG
2.2.3 ACCESSORIES
TRANSDUCERS
1. Disposable
2. Turbine
3. Fleisch
ACCESSORIES STANDARD
4. Thermal paper 110x50
5. USB and printer cable connection
6. Nose clip
7. Plastic Mouthpiece
8. Disposable mouthpiece pack (25 u)
9. Disposable transducers pack (25 u)
10. External power supply
11. Transducer holder
12. User’s manual
Easy Diagnostic
D T F D T F
1
511-B00-MU2 • REV. 1 .02
1
2.2.4 ACCESSORIES, OPTIONS AND
SPARE PARTS
REF.
OPTIONAL ACCESSORIES
Easy Diagnostic
D T F D T F
07046 Disposable
07053 Turbine
07052 Fleisch
OPTIONS
07272 Pulse oximetry module
FIRMWARE OPTIONS
07060 Bronchoconstriction module
07061 Concurrents display of V/t & F/V
07062 Database upgrade to 3000 records
07068
Diagnostic model Upgrade. Includes Whether station and bluettoh modules
ACCESORIES
02249 Calibration syringe (3 L)
02692 Nose clip (5 u)
01555 Cartoon mouthpiece 28x60 (100 u)
03169 Disposable transducers (25 u)
06391 SpO
2
soft sensor adult (M50B)
07725 SpO
2
soft sensor children (M50E)
02117 Antibacterial filter disposable
02118 Antibacterial filter reusable
02759 Antibacterial filter disposable membrane (50u)
07828 Software W20s CD
SPARE PARTS AND COMPONENTS
07233 Weather station module
07193 Bluetooth module
06610 Ethernet module
02634 Thermal paper 110x50 (5 u)
01145 Bluetooth adapter for PC
03658 USB 2.0 and printer cable connection
08165 Galvanic isolator for USB 2.0
511-B00-MU2 • REV. 1 .02
07147 Sniff probe (for MEP-MIP module)
01149 Paediatric Mouthpiece adapter
Included Optional
1
511-B00-MU2 • REV. 1 .02
2.3 LAYOUT OF CONTROLS AND CONNECTORS
FRONT PANNEL REAR PANNEL
1
BASE
Do not connect accessories not supplied by
SIBEL,S.A.
13 14 15 16
1- Internal printer (thermal paper roll).
2- Lever to lock/unlock the paper roll.
3- ON/OFF button
4- Color graphic touch screen (
640 x480 pixels).
5- Printer casing
6-Transducer Connection: Fleisch, turbine or Lilly disposable
7- MIP-MEP Connection 8- Pulse oximetry sensor connection
17 18
9- Power supply connection
10- USB connector for PC
11- USB connector for external printer
12- Ethernet Connection
13- Rechargeable battery location
14- Pulse oximetry board connection
15- Electronic Weather Station board connection
16- Bluetooth board connection
17- Lithium battery CR1815
18- Specifications plate
511-B00-MU2 • REV. 1 .02
1
2.4 INSTALLATION AND START-UP
This spirometer has been manufactured using solid-state professional components under strict quality controls. However, accidents may occur during the transportation or storage of the equipment and it is therefore wise to initially check its condition and that of its accessories before installing them.
If you detect any damage to the packaging, contact the haulier agency and distributor immediately before starting the installation.
Packaging must not be disposed of until the correct working order of the device has been fully verified.
2.4.1 START-UP
1
Connect the external power supply to the socket no.9, located on the rear of the device, and to the mains.
2
Insert the plug of the transducer cord into the socket no.6
3
Connect the other optional modules that you have acquired.
Consult the specific user’s manuals.
4
Press the ON/OFF button (no.3).
5
Choose the language
6
Enter the PIN number, if equipment protection is enabled. See the chapter
3.3 EQUIPMENT PROTECTION.
The first time the equipment is started, the protection is disabled and the PIN is set to 0000.
7
If equipment protection is not enabled or if the correct PIN has been entered, the
MAIN MENU
will then be displayed, which varies according to the model:
511-B00-MU2 • REV. 1 .02
1
Access to the FVC test
Access to the Post Bronchodilation test
Access to the VC test
Access to the MVV test
Access to the Bronchoconstriction test
Access to the Pulse oximetry test
Access to the MIP-MEP test
Quality control procedure: Calibration check
Access to the Internal database/ Database customization
Printing a report
Access to the Setup options: Calibration, Customization and Maintenance.
Pressing on the black area of the screen you may enable / disable the main menu buttons.
511-B00-MU2 • REV. 1 .02
1
2.4.2 RECHARGEABLE BATTERY: INSTALLATION AND LOAD
The DATOSPIR TOUCH spirometer operates optionally with
rechargeable Ni-Mh battery (10.8V 2500mAh) with an autonomy of 1.5 hours approx.
To install the battery, remove the base cover and insert it
Connect
where indicated in the figure.
The battery is recharged by connecting the spirometer to the power supply, even though the device is turned off. The charging time is about 20
hours.
DO NOT charge other type of batteries, they could
EXPLODE. Remove old batteries to avoid they could spill its substances.
2.4.3 POWER SAVING
To save power, when working with battery, the device includes an auto switch off system that turns the equipment off when the screen is not accessed for one minute, except in tests screens. In this case, you will be prompted to save the data before powering off.
When the spirometer is connected to the power supply, the backlight turns off after not accessing the screen for two minutes. The device will sleep until one key is pressed, recovering the information previously displayed.
511-B00-MU2 • REV. 1 .02
1
2.4.4 PLACEMENT OF THE PAPER INTO THE INTERNAL
PRINTER
Open the printer cover, lift the lever that unlocks the pull cylinder and insert the paper roll. Pull a small amount of paper out, put down the header lever, pass the paper through the slot of the cover and close it. A screen will appear to pull the paper in/out. Cut the paper pulling it forwards.
2.4.5 CONNECTION TO AN EXTERNAL PRINTER
Select the external printer option in the
COMMON
CUSTOMIZATION
menu. Connect the USB cable into the socket n.11
(PRINT) and the other end to the printer.
Contact the manufacturer or dealer for a list of compatible printers.
2.4.6 USB CONNECTION TO PC
To connect the device to a PC, install the USB driver and the
W20s Spirometry Software in the PC. Consult the W20s
Spirometry Software User’s Manual.
Then, connect the correct end of the USB cable to the socket n.10
, and the other end to the computer.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 22
2.4.7 BLUETOOTH CONNECTION TO PC
If you have purchased this module with the spirometer, it will be already installed. Otherwise, you must remove the cover from the base of the spirometer and insert the bluetooth board where is shown in Figure.
Then, in both cases, connect the
Bluetooth adapter to the PC and simply install the software included. To do so, consult the
Bluetooth adapter user’s manual.
Install the W20s Spirometry Software to the PC and choose the
Bluetooth link option, consult the
W20s Spirometry Software
User’s Manual
,
included in the CD.
The Bluetooth is enabled (and disabled) by accessing the Bluetooth option on the
COMMON CUSTOMIZATION
screen of the spirometer.
(See section
3.2.3
). Once the Bluetooth adapter has been configured in the PC, a connection will be established every time the equipment and the PC are started. The PC is then ready to receive the data transmitted by the equipment.
2.4.8 INSTALLATION OF THE WEATHER STATION
As in the previous case, if you have purchased the weather station module with the spirometer, it will be ready to use. Otherwise, remove the bottom case of the spirometer and insert the weather station’s
board, as shown in the figure.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 23
3. SPIROMETER CONFIGURATION
3.1 SETUP
Press and access to the options of the
SETUP MENU
:
1. CUSTOMIZATION
2. MAINTENANCE
3. CALIBRATION
3.2 CUSTOMIZATION
3.2.1 INITIAL CUSTOMIZATION
When the unit is turned on for the first time, the following screen will appear:
Select the desired language to operate with the spirometer, by pressing on the yellow square. This mark will appear on the option selected. Validate the selection by pressing .
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 24
3.2.2 TOOLBAR CUSTOMIZATION
It displays the values of atmospheric parameters:
Temperature, Pressure and Humidity. The temperature is automatically obtained from an internal sensor incorporated to the spirometer and its value will always appear in the tools bar. If you acquire the weather station module; besides temperature, will be shown the real values of pressure and humidity. Otherwise, touch this label to enter them, manually.
Touch on this label to choose the operating mode among Occupational Medicine, Primary Care or Diagnostic. (EASY:
OC, AP / DIAGNOSTIC: OC, AP DG).
The spirometer has been developed to make it user-friendly. Ease of use is assured through
context sensitive help screens that explain every DATOSPIR TOUCH feature, by touching this icon .
Touch this icon to adjust brightness.
Clock - calendar: Touch the displayed time to adjust time and date.
Power supply cord detection / Battery status: It displays the levels of battery charge by the segmented battery icon. When this icon turns red the battery is nearly discharged and it must be recharged.
See the battery’s charging procedure in section
2.4.2.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 25
Touch this icon to enter directly to the
SETUP MENU.
3.2.3 COMMON CUSTOMIZATION
If you wish to change any logging parameter, do so under this menu.
Access this option by pressing
, from the main screen, and then
.
The options included in the
CUSTOMIZATION MENU
are:
Default Configuration: This option memorizes a userdefined customization program status
Load configuration: restores the default customization.
Save configuration : Allows you to save the current configuration as the default.
Internal Database customization
•
Number of registers to be advanced if performing a fast advance using the database search engine
Save 3 maneuvers from the FVC test
•
•
Sort database
Common customization
Language
Enter a report header
Select the printer type: internal or external
Set clock-calendar: time and date
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 26
Set units: cm/Kg, in/lb, ºC/ºF or mmHg/hPa
Main menu icons (enable/disable)
Equipment Protection code (PIN)
Bluetooth: (enable/disable)
Ethernet: (enable/disable)
Initial customization: main menu icons, spirometry parameters, units, predicted, interpretation, incentive chart, pin, printer configuration and report header.
Spirometry customization
Select the Operating mode
Incentive for children and adults
Time audit: (enable/disable) show the time when the spirometry maneuver was performed
Predicted sets :
-
Choose predicted set among several authors for children and adults
-
Prioritizes the age range selected for adults if a different table is chosen for children.
-
Extrapolates the values for the ages outside the selected table range.
-
Ethnic factor
Observed parameters: (This is only at display level or for the report. All the parameters are saved on the database and can be enabled at any time).
Graphs and report customization:
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 27
-
Save graphs on the database
-
F/V- FVC Report
-
V/T- FVC Report
-
VC Report
-
MVV Report
-
D/R Report
-
Print large curves
-
Print 3 curves PRE
-
Print Data of 3 best maneuvers PRE
Interpretation: Select diagnosis according to
ATS/ERS, Miller, Snider/Kory/Lyons or NLHEP.
Printing the interpretation.
Quality control alerts:
-
Display of quality alerts to assure compliance with
ATS/ERS (EX, ET) or NLHEP criteria (QC Prompts)
Calibration check alerts:
-
Date of the last calibration
-
Require a daily Calibration check
-
Enable the use of the Spirometer without
Calibration check .
Bronchodilation test customization
Bronchoconstriction test customization (See the
Bronchoconstriction user’s manual)
Pulse oximetry customization (See the Pulse oximetry user’s manual)
MIP-MEP customization (Consult the MIP-MEP user’s manual)
Press to validate the values entered and to pass to the following screen, to cancel and to escapes the actual screen and to move back to the previous one.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 28
3.2.4 EQUIPMENT PROTECTION
In compliance with the data protection Act. Directive
95/46/EC, the DATOSPIR TOUCH has a protection option accessed by a PIN of 4 digits to prevent access to the equipment and, more specifically, to the private data it contains by unauthorized people.
Enable this option in the
CUSTOMIZATION MENU
(consult section
3.2.3
), by ticking the “PIN Enabled” checkbox and enter the PIN in the “New Pin” boxes. To change the PIN, you must enter the current one in the “Current Pin” box and then the new one in the “New Pin” boxes. The PIN can be disabled again by entering the current PIN and unticking the “PIN-enabled” checkbox.
If enabled, a screen will appear requesting the PIN, when the equipment is turned on:
ENTER PIN
Pin: 0000
Enter the PIN and you may access to the main menu screen. If an erroneous PIN is entered three times, the equipment will lock and switch off. On restarting it, a screen will appear requesting the
unlock code or PUK (supplied upon purchasing the equipment).
UNLOCK CODE (PUK)
000000000000000
If the correct code is entered, the device will unlock and the main screen will appear. From then on, the equipment will return to its initial status (Protection disabled and PIN 0000). If an erroneous code is entered, the equipment will remain locked.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 29
3.3 INTERNAL DATABASE
The DATOSPIR TOUCH has an internal database to store the tests performed for later review, print and / or transfer to a computer
(using the W20s Spirometry Software) or other computerized systems.
There are two storage capacities:
Database «L» (1000 tests)
Database «H» (3000 tests)
The database always saves all the spirometric parameters, despite them not being selected in from the
CUSTOMIZATION MENU
(Section
3.2.3
).
It is possible to save spirometric, Bronchoconstriction,
MIP-MEP or pulse oximetry test.
The base information remains, even when the equipment is turned off or the rechargeable battery is removed.
Press from the main menu:
Browse database: Display of the registers saved
Search patient: by ID code or last name.
In addition, you may:
• Display tests
• Print tests
• Delete tests
Delete database
Print summarized report of all the test saved into the database
Database customization:
• Number of records
• Save 1 or 3 maneuvers in tests, etc.
• Choose type of organization (ID or last name)
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 30
3.4 MAINTENANCE PROGRAM
The equipment has an internal diagnostic program to perform system checks and several adjustments.
DO NOT OPEN the instrument to carry out the maintenance program. Doing so, within the WARRANTY PERIOD WILL VOID THE
WARRANTY.
From the main screen, press
MENU
options are:
and then
.
The
MAINTEANCE
Equipment’s hardware Check-up:
CPU
ADCs check-up
Touch panel
Printer
Reindex the database
Reset all variables
ATS curves pre-saved: checkup of the product with pre-saved standard curves FVC, VC, MVV
Calibration and/or maintenance alerts.
Adjust of screen brightness and Internal printer
contrast.
Dealer’s data.
Equipment configuration: change from positive VC to negative VC or vice versa-
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 31
Adding a new transducer. Enter the pre-calibration code, delivered by the After-sales service with the new transducer, to activate it and to load the corresponding calibration factors.
Adding a new option. Enter the new Activation code delivered by the After-sales service to enable the new option purchased.
Display the current activation code.
3.5 QUALITY CONTROL: CALIBRATION CHECK
The
2005 ATS/ERS TASK FORCE
recommends that all spirometers be daily checked for calibration, before further testing begins.
The possible aging or the accumulated dirt of transducers may do an inaccurate measure. In the Pneumotachometers, the relationship between the pressure drop and the airflow depends on the gas viscosity. This viscosity is also depending of atmospheric conditions of temperature, pressure and humidity. This is the reason to carry out a calibration check every day to validate that
the device is within calibration limits.
If a device fails its
calibration check
, then new calibration procedure or equipment maintenance is required to ensure the proper use of the Spirometer.
To verify that transducer operates properly, the spirometer includes a simple check-up procedure based on measuring the known volume of a calibration syringe.
Test procedure:
1
Connect the spirometer to the 3 or 6L-syringe, as in the following figure, inserting a tube one meter long to avoid the
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 32
influence of the turbulence caused by the abrupt departure of the air:
DISPOSABLE
SYRINGE HOSE TRANSDUCER
CALIBRATION ADAPTER
CALIBRATION SYRINGE
1
Power up the DATOSPIR TOUCH and press from the main menu.
2
Discharge the syringe once at any range of the following flows varying between 0-12L/s:
Low flow level: 0 - 1 L/s
Mid-flow level: 2 - 5 L/s
High flow level: 6 - 12 L/s
To simplify the procedure, the spirometer allows you to carry out a single maneuver at Mid-flow level: 2 - 5 l / s.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 33
3
Press to cancel the maneuver or to accept it and go to the next range of flow.
4
At the end of the operation, press to see the results of the maneuvers performed: the percentages of variation and calibration factors calculated as the average of all the maneuvers performed on a range flow.
5
If the session is correct, the results will be automatically saved to the calibration record.
Press to see the data of the record (See section
3.7
).
6
Once calibration check performed satisfactorily, access the
Spirometry program to begin the tests.
7
If the calibration check is not correct, do a full calibration of the instrument. To do so, see
3.6 CALIBRATION PROCEDURE.
The error at each flow should be less than ±3.5%. The
expiratory and inspiratory factors remain unchanged in this routine.
You can program the spirometer, in the
COMMON
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 34
CUSTOMIZATION
menu, to require a daily calibration check before allowing access to any test. By default, this option is disabled.
3.6 CALIBRATION PROCEDURE
This procedure should be followed if the calibration check is not correct. The Calibration procedure is similar to the Calibration check, but in this case the calibration factors (expiratory and inspiratory) will be recalculated as the average of all maneuvers performed and applied on the spirometer. It will be necessary to deliver 3 maneuvers at the range flow of 0-12L/s.
To simplify the Calibration procedure, the spirometer allows you to carry out at least 3 maneuvers at Mid-flow level: 2- 5 l/s.
To carry out the calibration procedure, press and then .
Enter the necessary data described below and follow the instructions detailed in the section
3.6.
• Volume of the syringe: 1-6L.
• Humidity (%): set to 0.
• No. of Pulses of the
Turbine transducer
. Each turbine is factory calibrated individually and is associated to a factor
(3-digits) equivalent to the pulses/liter detected and printed on it.
• Factor of the
Lilly Disposable transducer.
Each pack of transducers is factory calibrated is associated to a factor printed on each lot. (3-digits).
If several sessions have been performed, only the correct last one will be saved. Then, you can print a report with the calibration record. If calibration is performed to more than one level of flow, the results on screen and on the report will be sorted from lowest to highest flow.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 35
3.7 CALIBRATION RECORD
The spirometer has a record containing the information of the last
30 calibrations check and calibrations performed, indicating: date, time, volume accuracy percentage and average flow for inspiratory and expiratory factors. This is extremely useful for centers requiring a quality control of the processes they use and also to help define day-to-day laboratory variability.
From the results screen:
Press to delete a record and to delete the entire database.
P ress the record desired.
to print the calibration report. Use to select
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 36
4. SPIROMETRIC TESTS PROCEDURES
The procedures to be completed to carry out the Forced Vital
Capacity «FVC», slow Vital Capacity «VC» and Maximum
Voluntary Ventilation «MVV» tests are very similar. Therefore, only one detailed description will be given in this section.
4.1 FORCED VITAL CAPACITY «FVC» TEST
1
Ensure that the transducer is plugged in the correct socket.
Insert the antibacterial filter and/or the disposable
mouthpiece into the transducer, as in the figures:
Lilly disposable
Fleisch and turbine
2
Turn on the spirometer and press on the Main Screen.
The first time per day, you carry out a spirometric test, the spirometer may advise you to check the calibration, as TAS-ERS recommends. Enable this option on the
SETUP MENU
.
4.1.1 ENTERING PATIENT DATA
Then, the following screens will appear
:
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 37
ID:
Name:
Surname:
Age:
Weight:
Height:
Sex:
Race:
Smoke yrs:
C/d:
Patient identifier, 10-character alphanumeric.
20-character alphanumeric field
25-character alphanumeric field
Number of years, between 2 and 100.
Between 15 and 200 kg / Between 33 and 440 Lb
Between 50 and 230 cm / Between 20 and 90 In
Male or female
Ethnic Factor, between 80 and 120%. It is used in areas without their own predicted equations, so data needs to be corrected to a specific percentage. 100% is equivalent to the unmodified standard value of the predicted.
If the NHANESIII predicted set has been chosen, in the
Customization menu, the RACE can be changed from the following groups: Caucasian, Afro-American and Hispanic.
Number of years the patient has been smoking
Cigarettes smoked per day. Between 0 and 100. This helps to calculate the COPD index parameter.
Technician
Transducer
Disposable
10-character alphanumerical technician’s code or name.
Transducer’s pre-calibration factor
(1)
Code:
(1)
The disposable transducer is pre-calibrated at the factory associating a calibration factor that have to be entered into the spirometer to measure correctly. (a unique factor for each lot of transducers).
The calibration factor can vary from lot to lot. Verify that the factor entered into the equipment matches the one of transducers lot that you are currently using.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 38
3
Enter the patient’s details by pressing on each field. An alphanumeric or numeric keyboard that will appear, depending on the field data type required. Enter patient’s data and press
.
If patient already exists in the database, press
to retrieve the patient’s details from memory. Once patient selected, the patient’s ID will appear at the top of the screen. When finish press to go to the test screen.
3
Instruct the patient on the test performance, his cooperation is essential for proper implementation.
The test must be carried out by medical staff. Review spirometry technique bibliography or request information about spirometry courses to SIBEL S.A.
4
When the patient is ready and have assumed a correct posture, fit the nose clip. The subject must hold the transducer, without moving it, and wait until a blinking arrow appears on the screen.
The transducer must remain in the same position until the
end of the maneuver.
Incorrect
Correct
5
Patients can carry out the spirometry maneuvers in either of two different ways:
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 39
• Starting the maneuver with FORCED
EXPIRATION
followed by FORCED
INSPIRATION, if desired.
• The patient breathes normally and when indicated by the technician, takes a deep breath completely filling his lungs, and then performs a FORCED EXPIRATION followed by FORCED INSPIRATION, if desired.
6
The device will detect the end of maneuver as ATS/ERS criteria and then will display the resulting graph and the parameters.
The maneuver performance can be ended at any time by pressing
, which will appear on the bottom left-hand of the screen.
At the end of the maneuver, one or more of the QC Alerts (if the option is enabled in the
Customization menu
) may appear, alerting the technician as to whether the maneuver is in compliance with ATS/ERS or NLHEP Quality Control criteria. Consult the section
4.2.
ATS/ERS Criteria: The maneuver with least warnings is considered the best (ET, EX). With the same number of
ATS/ERS warnings (ET, EX), the maneuver with the highest
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 40
sum of FVC+FEV1 is considered the best.
7
Perform at least 3 maneuvers, but no more than 8, as this would tire the patient.
If 3 or more maneuvers have been performed and the FVC and/or FEV
1 parameters blink on and off, this indicates that the
repeatability criteria has been fulfilled for one or both parameters, according to the ATS/ERS criteria. This criteria indicates that the two best FVC or FEV
1
values do not differ by
more than 150 ml if FVC is more than 1 liter or more than
100 ml if FVC is less than or equal to one liter.
4.1.2 DISPLAYING RESULTS AND OPTIONS
•
•
At the end of the test:
The current, predicted and best maneuver graphs are shown.
On the summary frame, the FVC and FEV
1
values of all the maneuvers performed will appear:
PRED: indicates the patient’s Predicted value
*: indicates the current maneuver
Mx: indicates the selected maneuver
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 41
Maneuvers are sorted from best (M1) to worst (M8) according to
ATS/ERS criteria and are shown in different colors: green for those that are acceptable and repetitive and from yellow to red the lower-quality maneuvers. In white, the current maneuver.
•
Press to enlarge or reduce the graph.
•
Press on the plot to switch from one graphic type to the other: Flow/Volume, Volume/Time, both or incentive and a small Flow/Volume graph.
• Selection of the desired maneuver Mx :
the scroll allows you to move over the maneuvers performed and select the desired one to see the graph, view the parameters, consult the
diagnosis, delete or save a maneuver or print out a report.
(Consult the sections from 4.1.4 to 4.1.9)
By default, the maneuver selected is the best, M1.
• Press to change the
Incentive chart: a Temporal bar:
Progress bar in function of time, in three colors: red indicating less than 4 seconds, orange between 4 and 6 seconds and green for more than 6 seconds.
b Volume bar:
Progress bar in terms of the expired volume.
Red indicates 75% below the predicted value, the orange between 75% and 100% and green 100% above the predicted value.
c Incentives for children
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 42
• Time audit:
show the exact time of testing have been carried out in the reports for a further monitoring.
•
Press on ID area to access the patient’s data Screen.
•
Press the performed.
to display all parameters data of maneuvers
• Observed values of the parameters selected in the
CUSTOMIZATION MENU .
• Predicted values <REF> (If patient’s data has been filled in)
• % between both parameters. If an * appears after the REF test, this means that the predicted values have been extrapolated
• The best FVC and FEV1 values (not necessary from the same maneuver).
• QC grade
• ATS/ERS alerts
• Time audit.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 43
The screen only displays the results of 3 maneuvers and the predicted. By default, it appears the data of the 3 best maneuvers:
, and . To view the data of the other maneuvers performed, press on its button , , , or .
The BEST maneuver is set at M1. Therefore, this will be used to display the interpretation, to print the report or to save the maneuver. The technician has the option to override this selection by pressing , . , , , or .
4.1.3 DELETING AN EXISTING MANEUVER
Select the maneuver that you want to delete and press
4.1.4 SAVING A TEST TO THE INTERNAL DATABASE
1
Select the maneuver you wish to save. The maneuver selected by default is the best (M1).
2
Press to save the best maneuver or the 3 best
maneuvers on the database, according to the option selected in
SETUP MENU
(Option not available in OC mode).
4.1.5 VIEWING THE INTERPRETATION
Press to view the interpretation According to the diagnosis available on your device’s operating mode or depending on the customized option selected on the
CONFIGURATION MENU
:
Miller - ATS/ERS - Snider, Kory & Lyons or NLHEP.
4.1.6 PRINTING A MANEUVER
Print the test report via the internal or external printer. Select the desired printer on the
CUSTOMIZATION menu.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 44
Then, select the maneuver to be printed and press to print the report of the maneuver selected and to print the report of the 3 best maneuvers, if the “Save 3 Maneuvers” option is activated.
The report will include the parameters and graphs corresponding to the selected maneuver o 3 maneuvers. If you do not want graphs, certain parameters, the diagnosis and/or ATS/ERS warnings to appear, disable them on the
CUSTOMIZATION MENU.
Moreover, by pressing in the Main menu you may print:
The last test performed
All the tests carried out during the day
Whatever test you want of the database
4.1.7 OTHER TESTS ON THE SAME PATIENT
After carrying out the FVC test on a patient, you may do the following:
• A VC test on the same patient
• An MVV test on the same patient
• A Post bronchodilator test on the same patient
• Enter data for a new patient.
• Print the general report of all tests on the same patient
The spirometer saves the best maneuver of each FVC, VC,
MVV and/or bronchodilator test to print a general report.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 45
4.1.8 CHANGING PATIENT DATA/ ENTER A NEW PATIENT:
This option allows you to enter a new patient data or to modify existing patient data. (Follow the procedure described in the section
ENTERING PATIENT PARAMETERS
).
You may access to this menu by pressing on the ID area from the tool bar in any test.
If you have modified existing patient data (age, height, sex, etc.) the predicted parameters will be calculated again.
4.2 QUALITY OF FVC TEST
The DATOSPIR TOUCH spirometer includes quality alerts to assist you in producing high quality spirometry tests. These alerts may be enabled or disabled in the
CUSTOMIZATION MENU.
I. ATS/ERS ALERTS
To ensure good spirometry, the technician will pay particular attention to ensure that the patient has made the utmost effort, that the start has been good and that no coughing or Valsava’s maneuver due to glottis closure has occurred. Special attention must be paid to preventing expiration from ending too soon.
Once the maneuver is finished, one or two Alerts may appear on the screen indicating that the maneuver has not been performed according to the ATS/ERS criteria:
ET
- This indicates that expiration did not end satisfactorily because the variation of accumulated volume during the last second of the maneuver was over 25 ml, or that the maneuver lasted for less than 6 seconds (in individuals 10 or older) or less than 3 seconds (in children under 10).
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 46
EX
- This indicates that the expiration did not start
satisfactorily because the extrapolated volume was greater than 5% of the FVC or 0.15 liters. According to the ATS/ERS recommendations, this volume should be lower than 5% of the
FVC or 0.15 liters, whichever is greater.
The technician performing the spirometry may disable these alerts in the
CUSTOMIZATION MENU.
In this case, they will also be removed from the printed report. This disabling is only at display level. The warnings are still taken into account when classifying the order of the maneuvers.
II. NLHEP ALERTS: QC PROMPTS
In order to assess the pulmonary function of the patient, it is necessary to obtain acceptable test quality. The test quality depends on cooperation of the patient and this, in turn, depends on the quality of the technician's instructions.
Accordingly, DATOSPIR TOUCH incorporates an automatic quality control function, based on the recommendations of the National
Lung Health Education Program (NLHEP), with prompts to assist the technician in providing the good instructions to the patient to produce high quality spirometry tests.
At the end of a maneuver, a message on the screen will inform you as to whether the maneuver was acceptable or not. If not, a prompt will guide you on how to coach the patient to do it better.
(See the rows in white of the following table).
As soon as the message "Good test Session" appears, do not carry out further maneuvers. If, even after repeated attempts, it is not possible to obtain an adequate number of good maneuvers, you
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 47
should take a break, depending on how the patient feels or stop measurement.
Only one of the following QC prompts is displayed after
a performed maneuver (in the order of priority listed below).
QC Prompt Criteria How to improve the maneuver?
Don't Hesitate
Blast Out
Faster
Blow Out
Longer
Blast Out
Harder
EX error
Time to PEF higher than
120ms
ET error
If there are not 2 acceptable maneuvers, with at least the largest 2 PEF values matching within 1 L/s
The patient must start exhaling harder.
The patient must exhale as hard, firm and fast as possible.
The patient has abruptly interrupted exhalation. The patient must exhale even more and expel as much air as possible from his/her lungs.
The maneuver differs significantly from the previous ones The patient can exhale even more vigorously and achieve a higher peak flow.
Deeper Breath
If there are not 2 acceptable maneuvers, with at least the largest 2 FVC values matching within 150mL and 1L/s for
PEF.
The maneuver differs significantly from the previous ones. The patient must inhale more deeply and exhale even more air.
Good Test
Session
After 2 acceptable maneuvers with at least the best 2 maneuvers match.
TEST COMPLETE. Adequate number of good maneuvers.
Referring to the quality of the last maneuver performed
Referring to the reproducibility of the maneuvers performed
III. NLHEP QUALITY GRADING (QC GRADES)
At the end of the test (maneuvers session), a quality grading from
A to F will be displayed to indicate the reliability of the results, according to NLHEP criteria.
A, B and C grades indicate a reliable result, but a grade D or F indicates a poor quality test (in this case, the results should be interpreted with caution).
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 48
GRADE TEST CRITERIA
A
B
C
D
VERY GOOD
GOOD
ACCEPTABLE
POOR
At least 2 acceptable maneuvers with the largest 2 FEV values matching within 100mL and the largest 2 FEV than 100mL.
1
6 values matching better
At least 2 acceptable maneuvers with FEV
1 values matching between 101 and 150mL
At least 2 acceptable maneuvers with FEV
1 values matching between 151 and 200 mL
Only one acceptable maneuver, or more than one, but the FEV
1 no interpretation)
values match > 200 ml (with
F
NOT ACCEPTABLE
No acceptable maneuvers (with no interpretation)
4.3 POST BRONCHODILATATION TEST
The DATOSPIR TOUCH spirometer allows spirometry tests: FVC,
VC or MVV after the administration of a bronchodilator drug. For this, the tests should have previously been completed in PRE bronchodilator mode and saved to the PRE database.
The procedure to carry out POST-Bronchodilator spirometry is:
1
Complete an FVC, VC or MVV test to the patient as described in the
4.1
section before administration of the bronchodilating drug.
2
Save the PRE test in the database for comparison to the
POST-Drug test.
3
Administer the dose of the bronchodilator drug prescribed and wait for the standardized period.
4
On the main screen, press
.
A screen similar to the following will appear which displays the tests saved in PRE mode.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 49
5
Select the PRE test with which to be compared and press .
(FVC selected by default).
6
Using keys and , the VC and MVV tests can be seen and saved in PRE mode.
The screen then shows the two graphs (PRE and POST) for comparison purposes:
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 50
The data screen shows the observed values in PRE and POST mode and the method of comparison between them, depending on the option selected in the
CUSTOMIZATION MENU
.
• Weighted % between PRE and POST
• % between REF (Predicted) and POST
• % between PRE and POST
• Difference between PRE and POST
Thus, it may provide a single report of spirometric results before
(PRE) and after (POST) application of a bronchodilator drug.
4.4 SLOW VITAL CAPACITY «VC» TEST
1
Access the test by pressing , from the main screen.
•
The axes are always displayed in VOLUME/TIME mode.
•
The equipment saves a maximum of 8 maneuvers ordered according to the VC value, where M1 is the best VC and M8 the worst.
•
The maximum time allowed for the maneuver is 45 seconds.
•
To measure the ERV and TV parameters correctly, each maneuver must have at least four respiratory cycles.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 51
4.5 MAXIMUM VOLUNTARY VENTILATION «MVV» TEST
1
Press , from the main screen.
2
Inhale and exhale completely without interruption for at least 12 seconds.
•
The axes are displayed in VOLUME/TIME mode.
•
The maximum time allowed for the maneuver is 15 seconds.
•
The equipment saves a maximum of 8 maneuvers, ordered according to the MVV value, where M1 is the best MVV and M8 the worst.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 52
5. COMMUNICATIONS SYSTEM
One of the strengths of the DATOSPIR TOUCH spirometer is its
Communications System, which allows the user to:
1. Transfer patient tests to a PC
2. Transfer Equipment Checking Data
3. Update Internal Software
4. Export patient tests to other Management Systems
Communications can be made via USB (standard), Bluetooth
(optional) or Ethernet (optional).
5.1 TRANSFERRING TESTS TO THE PC
If you want to view, print, manage and/or save the tests to the PC, you must have W20s Spirometry Software.
The process to follow is:
1
Save the tests required in the equipment’s internal Database
2
Install the W20s Spirometry Software, as detailed in its
User’s manual.
3
Load the data to the PC, by pressing on the W20s
Software. The screen shows a list of the tests transferred and you can select those to be imported to the PC Database selected in the W20s Software SETUP option. For USB compatible spirometers, a driver must be installed on the PC.
4
Then you can select, view or print any of the tests imported or transferred to the PC.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 53
5.2 TRANSFERRING EQUIPMENT CHECK DATA
The DATOSPIR TOUCH includes a program that auto-checks the working order of certain parts of the equipment, displaying the information on the screen and saving it in an internal file.
• Hardware checkup
• Firmware checkup
• Equipment customization
• Calibration Record
• FVC test with standard curves
1
Start up the spirometer, press from the main menu and select
MANITENANCE
. Access the Equipment Check option and run all the sub options, following the instructions on the screen.
2
Connect the equipment and the PC via USB or Bluetooth.
3
Run the previously installed W20s Spirometry Software, making sure that “DATOSPIR TOUCH” is selected in
Configuration - Links and access the Configuration -
Utilities - Download Data option.
The transferred information is saved in the DATA directory of the application, in the files:
STATUS.CSV
Contains the errors detected
CALIBRA.CSV
Contains the calibration data
CONFIG.CSV
Contains the equipment customization
PRUEBAS.CSV
Contains the database tests
GRAFXxx.CSV
Contains the graphs in Flow/Time mode
The files from the previous transfer are renamed with the extension .OLD
4
If you want to view the information of any of the files, load them using EXCEL.
5
If a problem is detected that the user is unable to solve, send the
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 54
auto-check information by e-mail or fax to the SIBEL S.A.
After-sales Service or to your distributor, who will analyze it and assess the cause of the problem, providing or proposing a suitable solution.
5.3 ADDING OPTIONS AND/OR TRANSDUCER
If you want to add a module / option (SpO
2
, MIP-MEP, bluetooth, weather station, bronchoconstriction, etc.) and/or a new transducer; please contact the After-sales Service who will send you the necessary components and the activation code.
Enter this code to the device by pressing on , from the main screen, and then
(MAINTEANCE MENU)
.
Press to add a new transducer and / or press new module / option. to activate a
In both cases, the following screen will appear, where you must enter the activation code that will be delivered with the module or transducer.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 55
5.4 UPDATING FIRMWARE
The Datospir TOUCH contains internal software (firmware). SIBEL,
SA carries out improvements on it continuously. If you want to incorporate them into your spirometer and use the last version of the program, contact
After-sales Service
. This will send you some files to be stored in the same location and then follow these instructions:
1
Switch the equipment on.
2
Run the W20s Spirometry Software (the W20s in demo mode provided, upon purchasing the equipment, is enough).
3
Access the Configuration - Links option. Then, check that the
DATOSPIR TOUCH is selected. Access the Configuration -
Hardware Test option. Run a communications test to check that the connections are correct.
4
Access the Configuration - Utilities - Update Flash option. A dialogue box will open where the update key (provided by SIBEL) must be entered. Select the
Load.prj
file.
5
The new program will be transmitted. The process may take around 10 minutes, depending on the PC.
6
Switch off the DATOSPIR TOUCH.
7
Once loaded the program,
turn on the spirometer holding pressed the ON / OFF button for 5 seconds.
For information relating to the W20s Spirometry Software, see its
User’s Manual.
5.5 EXPORTING TESTS TO OTHER SYSTEMS
The DATOSPIR TOUCH spirometer can export the tests saved previously in the Internal Database to other management systems at other locations.
The equipment shows the information in comma-delimited
mode, making it compatible with many different systems.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 56
The information is available in the following files:
PRUEBAS. CSV
Contains the database tests
PATIENTS.CSV
Contains the database patients
GRAFxx. CSV
Contains the graphs in Flow/Time mode
The graph file, as indicated, contains the graphs for each test in
Flow/Time mode. If you want to display the graphs in Volume/
Time or Flow/Volume mode in the new management system, the following aspects must be taken into account:
• The Flow signal is sampled at 100Hz.
• The ratio of the axes in the Volume/Time graph must be
• adjusted to 1 liter = 2 seconds.
• The ratio of the axes in the Flow/Volume graph must be adjusted to 2 l/s = 1 l.
In the event of doubt or queries, contact the SIBEL S.A.
Technical Support or your distributor, who will provide any further information you may require.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 57
6. TECHNICAL SPECIFICATIONS
6.1 GENERAL SPECIFICATIONS
Power supply
Compatible external power supplies
Input 100 to 240V, 50 to 60Hz
Output 12V 2.5 A (Electrical Protection: Class I)
• MEANWELL MES30A-3P1J
• EMERSON DP4012N3M
• DANUBE FRM030-S124
Class IIa
Medical device classification
Protection level
IPX2: Device (not power supply) is protected against vertically dripping water when the enclosure is tilted up to 15 degrees
Dimensions and Weight
Storage capacity
(Database)
Communications
195 x 270 x 100mm
1,7 kg (with battery) approx.
• 2 types of database according to their storage capacity:
«L»:Small DB: >1000 tests including F/V loops and V/t curves.
«H»:Large DB: >3000 tests including F/V loops and V/t curves.
• Storage of spirometry, pulse oximetry and MIP-MEP tests.
• PC: USB 2.0, Bluetooth 2.0 (op.), Ethernet (op.)
• Printer: USB 2.0 Bluetooth 2.0 (op.)
Printer protocol
Display
Battery pack
Operating Conditions
Recomended
Conditions of measurement
PCL 3 HPA, PCL 5e high resolution touch screen LCD, VGA of 640x480
Ni-Mh Rechargeable, 9.6V 2.5Ah
Temperature: 5 to 40 ºC
Humidity < 85% (without condensation)
Pressure: 850 to 1060 hPa. (638 a 795 mmmHg/1500 a 0 m aprox.)
Temp
Peak Flow
(EN ISO 23747:2007)
10-35 ºC
Forced expiratory volume
(EN ISO 23747:2007)
17-35 ºC
ATS/ERS
Take Force
>17ºC
Hum 30-75% (sin condesación)
Internal temperature sensor
Press
Transport and storage
Temperature: -20 to 70 ºC
Humidity < 85% (without condensation)
5 to 40 ºC ± 1 ºC
850-1060 hPa
(638 a 795 mmHg / 1500 a Om aprox.)
Max. Nº of maneuvers per subject
8 FVC maneuvers, 8 VC maneuvers, 8 MVV maneuvers
Lifetime
Weather station
7 years
Pressure: 500 to 1040 hPa, accuracy: ± 6.67 hPa
(375 to 780mmHg, accuracy: ± 5 mmHg)
Humidity: 0 to 100% (accuracy: ± 5%)
Applicable standards
• European directive concerning medical devices 93/42/EEC (RD
1591:2009)
• Quality (EN ISO 13485:2003/AC:2009, EN ISO
9001:2008 and EN ISO 14971:2009)
• Compliance with data protection Act. Directive 95/46/EC
• Safety Medical devices (EN 60601-1:2006)
• Electro-magnetic Compatibility (EN 60601-1.2:2007/AC:2010)
• Biocompatibility: Biological evaluation of medical devices.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 58
Spirometry standards in force
(EN ISO 10993.1: 2009 / AC: 2010)
• Usability (EN 60601-1-6:2007)
• Aptitude of use (EN 62366:2008)
• Spirometers for forced spiratory volumes measurement
(EN ISO 26782:2009/AC:2009)
• Spirometers for peak spiratory flow (EN ISO
23747:2009)
• Software of medical devices (EN 62304:2006/AC:2008)
• Pulsioximetry (EN ISO 9919:2009)
• Vibration and temperature: (Series EN 60721:1995 and
Series EN 60068:1999)
• Documentation and information (EN 1041:2008, EN
980:2008)
• Waste disposal according to WAEE Directive 2002/96/CE.
• Water protection, IP (IEC 60529:1989+M1:1999 Ed 2.1)
• ATS/ ERS Standards:
1. Miller MR, Crapo R, Hankinson J, et al. General considerations for lung function testing. Eur Respir J 2005;
26:153–161.
2. Miller MR, Hankinson J, Brusasco V, et al. Standardisation of spirometry. Eur Respir J .2005; 26: 319–338.
3. V. Brusasco, R. Crapo and G. Viegi. Standardisation of the measurement of lung volumen Eur Respir J 2005; 26:
511-522
• SEPAR: Sanchis et al Normativa para la espirometría forzada.
Recomendaciones SEPAR núm. 1. Arch Bronconeumol
1989; 25: 132-142.
• NLHEP: Ferguson et al. Office Spirometry for Lung
Health Assessment in Adults. Chest 2000; 117:1146-
1161.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 59
6.2 SYMBOLS
SERIAL NUMBER
MANUFACTURER’ (Manufacturing date, manufacturer’s name and address)
LOT NUMBER
PRODUCT REFERENCE
EXPIRATION DATE
DO NOT REUSE
TEMPERATURE LIMITATION
CONSULT THE INSTRUCTIONS FOR USE
PRECAUTION
START-UP (STANDBY)
BF APPLIED PART
NO PHISIOLOGICAL PULSE OXIMETRY ALARMS
IPX2
WASTE DISPOSAL ACCORDING THE WAEE DIRECTIVE
SENSITIVE CONNECTOR TO ELECTROSTATIC DISCHARGES (See Annex 1)
ENTRANCE (EXTERNAL POWER SUPPLY)
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 60
6.3 TESTS, FUNCTIONS AND PARAMETERS
Available Information on tests:
• Percentage deviation in relation to predicted values.
• Standardized values of predicted that can be selected from several standards.
• Patient’s ID details.
• Atmospheric data on temperature, pressure and relative humidity.
• Saving of 8 maneuvers from the same study.
• Graphs in flow/volume and volume/time for FVC,
Bronchodilatation and Bronchoconstriction tests.
• Graphs in flow/volume for VC and MVV tests.
FORCED VITAL CAPACITY FVC
• FVC (l) Forced Vital Capacity
• FEV0.5 (l) Forced Expiratory Volume in 0.5seconds
• FEV1 (l) Same in 1 second
• FEV3 (l) Same in 3 seconds
• FEV.5/FVC (%) Ratio
• FEV1/FVC (%) Ratio
• FEV3/FVC (%) Ratio
• FEV1/VC (%) Ratio
• PEF (l/s) Peak Expiratory Flow
• PEFT (s) Peak Expiratory Flow
• FEF25% (l/s) Forced Expiratory Flow 25% into the maneuver
• FEF50% (l/s) Same, 50% into the maneuver
• FEF75% (l/s) Same, 75% into the maneuver
• FEF25-75% (l/s) Mean expiratory flow between 25% and 75% of the FVC
• FEF75-85% (l/s) Mean flow between 75-85% of FVC
• FET25-75 (s) Forced expiratory time between 25-75% of FVC
• FET100 (s) Forced Expiratory Time
•
MEF50/MIF50
(-) Ratio
• FEV1/FEV.5 (-) Ratio
• FEV1/PEF (-) Ratio
• MIF50% (l/s) Maximum Inspiratory flow with 50% of FVC inspired
• FIVC (l) Forced Inspiratory Vital Capacity
• FIV1 (l) Forced Inspiratory Volume in 1 second
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 61
• FIV1/FIVC (%) Ratio
• FEV1/FIV1 (%) Ratio
• PIF (l/s) Inspiratory Flow Apex
• MTT (s) Mean Transit Time
• PEF/PIF (-) Ratio
• Vext (%) Extrapolated Volume
• MVVInd (l/min) Maximum Voluntary Ventilation (30 x FEV1)
• FEV6 (l) Forced Expiratory Volume in 6seconds
• FEV1/FEV6 (%) Ratio
• EPOC rate Parameter that depends on the number of cigarettes smoked a day, the age and FEV1. Indicates the risk of EPOC.
• Lung Age Parameter that depends on the height and FEV1. This indicates the equivalent age of the lung.
• Quality alerts to assure compliance with ATS/ERS and NLHEP criteria
• Acoustic and graphic indication of the start and end of each maneuver
• Start FVC expiration: Using the retrograde extrapolation method
• End FVC expiration: When the volume accumulated in the last second is below 25ml.
BRONCHODILATATION TEST
• Same parameters and characteristics as in FVC.
• Several methods of comparison among PRE, POST and REF values.
• Superimposing of PRE and POST graphs.
SLOW VITAL CAPACITY
• VC (l) Slow vital capacity
• TV (l) Tidal volume
• ERV (l) Expiratory Residual Volume
• IRV (l) Inspiratory Residual Volume
• IC (l) Inspiratory Capacity
• Ti (s) Inspiratory time
• Te (s) Expiratory time
• Tt (s) Total time
• Ti/Tt (-) Ratio
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 62
MAXIMUM VOLUNTARY VENTILATION
• MVV (l/min) Maximum Voluntary Ventilation
• Br./min (Br/min) Breathing frequency of MVV
6.4 PREDICTED SETS
The spirometer includes several predicted tables that the user can select in the
CUSTOMIZATION MENU\ Spirometry
.
Predicted set
SEPAR
ERS
KNUDSON
CRAPO
ZAPLETAL
MORRIS
AUSTRIA
GUTIERREZ
CASTRO - PEREIRA
POLGAR - WENG
HANKINSON - NHANES III EEUU 4
PEREZ - PADILLA
CRUZ-MORALES
GOLSHAN
GARCIA RIO
CANDELA
Contry /
Region
SPAIN
EUROPE
EEUU
EEUU
EUROPE
EEUU
AUSTRIA
XILE
BRAZIL
MEXICO
MEXICO
Age range
(years)
6 to 70
(1)
18 to 70
(1)
6 to 84
(1)
4 to 91
(1)
4 to 17
24 to 100
6 to 90
(1)
5 to 100
(1)
6 to 76
(1)
IRAN
EUROPE 65
4
100
7 to 100
(1)
17 to 64
(1)
6 to 81
(1)
85
SPAIN 2
(1)
If other ages are used, the predicted values are extrapolated
6.5 FVC INTERPRETATION (Diagnosis)
• Miller chart
• Snider, Kory & Lyons
• Interpretation ATS/ERS. Pellegrino et al. Task force:
Standardisation of Lung Function Testing. Eur Respir J 2005;
26: 948–968
• Interpretation NLHEP. (Only valid for predicted values that calculates the LLN. For example: Hankinson). Ferguson et al.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 63
Office Spirometry for Lung Health Assessment in Adults. Chest
2000; 117: 1146-1161.
6.6 TRANSDUCERS
RANGES AND MEASUREMENTS (According to ATS/ERS2005)
Measurement Scale (BTPS)
Flow
Volume
Dynamic flow resistance
Precision of measurements (BPTS)
Volume
Flow
PEF
Time related precision
Volume resolution
Sampling frequency
Transducer lifetime
Fleisch Turbine Lilly
Dispos.
0 to ±16 l/s
0 to 10 l
< 1,47 hPa (1.5 cmH
2
0) / (l/s) a 14 l/s
(the highest value)
3% or 50 ml
5% or 200 ml/s
10% or 300 ml/s
<10ml
0,50%
<10ml <10ml
100Hz 100Hz 100Hz
1400 disinfect. or 7 years
600 disinfect. or 3 years
Only one use (Expir.
3 years)
6.7 MANUFACTURER’S LIABILITY
SIBEL S.A. is only liable for the safety, reliability and working order of this equipment if:
• The place where the equipment is installed or used is compliant with the requirements related to the IEC electrical installation and other applicable regulations.
• All repairs, services and modifications inside and outside the guarantee period are carried out by SIBEL S.A. technical staff.
• The equipment is used by skilled staff according to the recommendations of this User’s Manual.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 64
7.
CLEANING AND MAINTENANCE
DATOSPIR TOUCH spirometer requires cleaning and maintenance aimed at keeping the equipment functioning correctly and at ensuring safety of patients and operators.
7.1 CLEANING / DISINFECTION
The person who carries them out does not require any special technical knowledge other than their own understanding of the functioning and handling of the equipment. It is usually done by the normal user of the equipment.
For your safety, disconnect the equipment from electrical power or remove the battery before cleaning.
7.1.1 CLEANING / DISINFECTION OF THE TRANSDUCER
The transducer is the part exposed directly to the patient, so it must be kept in perfect physical and hygienic condition. To do so, proceed as follows:
A)FLEISCH PNEUMOTACHOMETER
1
Remove the filter and after the transducer by pressing slightly its tab so that is comes away from its
2
fixtures. See the figure.
2
Wash the transducer and the filter with soapy water. After cleaning, they should be rinsed briefly in distilled water.
Do not use abrasive substances or solvents.
1
3
Leave it to dry at room temperature and re-assemble the group again.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 65
B) TURBINE TRANSDUCER
1
Remove the turbine from its housing by pressing slightly its tab so that is comes away from its fixtures.
2
Wash the turbine with water and soap, avoiding
solvents or abrasive substances that may damage the components. Given that its reliability depends on the condition of the turbine, examine it for external damaged. Do not rinse the turbine by holding it under
running water.
3
Rinse briefly in distilled water to avoid salt deposits.
4
Leave it to dry at room temperature and re-assemble the turbine in the housing.
A/B TOP LEVEL DISINFECTION OF THE TRANSDUCER
If you suspect microbial contamination, replace Step
2
by using
CIDEX® OPA
solution (follow the manufacturer’s instructions).
C) LILLY DISPOSABLE PNEUMOTACHOMETER
This transducer does not need any kind of cleaning as they are only for single use. They must be thrown away once the patient has carried out the tests.
Reutilization of the disposables transducers gives a risk of
crossed infection. The use of disinfectant products can affect the transducer’s mesh and give a loss of accuracy in the measurement.
D) HANDLE AND TRANSDUCERS HOUSING
Clean internal and external parts of the transducer’s housing and handle with a dry or slightly water-moistened cloth, drying
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 66
thereafter the remanent humidity. Pay attention to avoid the introduction of strange elements in the pressure inlets of the Lilly handle. Do not use abrasive substances or solvents.
7.1.2. CLEANING THE SPIROMETER
The spirometer should be cleaned gently with a cloth moistened with water and dishwasher detergent. It can then be wiped dry.
Take particular care to ensure that no liquid enters the interior of the device or the connectors and connections.
Do not use abrasive substances or solvents. Remove the
batteries from their compartment before cleaning.
7.2 PREVENTIVE MAINTENANCE
Preventive maintenance consists of any actions aimed at keeping the equipment in a good working order.
Actions which can be carried out by the same user:
1
Each time the spirometer is turned on, the equipment will check certain parts and/or functions. In addition, the user can access the Maintenance Program to adjust and/or check any parts of the equipment, as indicated in detail in the corresponding section.
2
A second procedure, which can be performed by the user, consists of regular monitoring of the appearance of the different connections and other external parts of the equipment. Check that all connections are perfectly connected, that no cable and/or connector or any other element is broken or damaged.
3
Execute a calibration check with a 3-L syringe. Calibration checks must be undertaken daily.
4
Define the periods in days between calibrations or preventive maintenance work on the
CUSTOMIZATION MENU
of the equipment. If the days specified are exceeded, the equipment warns of such by displaying a sign every time it is started. If
“0” days is entered, a warning is never given.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 67
In the event of detecting any problem that the user cannot solve, contact the SIBEL S.A. After-Sales Service or your distributor to review or repair it.
Actions carried out by skilled technical personnel:
According to the different regulations, particularly the 93/42/EEC
Medical Device Directive, electromedical devices should be verified and/or calibrated regularly to ensure reliable functions and the safety of patients, users and the environment.
This Technical check will be performed every year following the
DATOSPIR TOUCH Verification and Adjustment Procedures, available from the manufacturer SIBEL, S.A. This type of operation must be carried out by skilled technical staff from the distributor’s or manufacturer’s technical service.
On all accounts, SIBEL S.A., as the manufacturer, must provide written authorization, for at least the guarantee period, for the corresponding technical personnel to carry out said maintenance and will not be held liable under any circumstances for any damage, malfunction, etc. that may arise as a result of defective maintenance by people not employed by SIBEL S.A.
7.3 CORRECTIVE MAINTENANCE
Corrective maintenance consists of repairing the equipment that has stopped working, due to malfunctioning or misuse; leaving it in a good state.
On detecting any fault with equipment which interferes with its normal use, disconnect the equipment from the mains and contact SIBEL S.A. After-Sales Service. Specify the problem in as much detail as possible.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 68
Annex 1.
ELECTROMAGNETIC COMPATIBILITY
1.1 ELECTROSTATIC DISCHARGE
The Datospir Touch uses the exception of electrostatic discharge test on the Ethernet connector as described in EN60601-1-2: 2007.
For this reason, as stated in paragraph 5.1.2 of the standard, the symbol must be placed next to the Ethernet connector.
This symbol does not imply any problem of safety to the technician or to the patient. Neither implies that the Ethernet module is damaged. It would only cause that the Ethernet communication stops running for a moment.
However, to avoid any slightest problem, you should take the following precautions:
- Connect and disconnect the Ethernet cable when the device is turned off.
- During the functioning of the equipment, do not touch the connector with your fingers.
- Inform to all staff involved of the meaning of the symbol and the precautions about discharge sensitivity.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 69
1.2 GUIDANCE & MANUFACTURER’S DECLARATION
Guidance and manufacturer’s declaration – electromagnetic emissions
DATOSPIR TOUCH is intended for use in the electromagnetic environment specified below. The customer or the user should assure that it is used in such an environment.
Emissions test
RF (Radiated) emissions
CISPR 11 (EN 55011)
RF (Conducted) emissions
CISPR 11 (EN 55011)
Complian ce
Group 1
Class B
Group 1
Class B
Class A
Electromagnetic environment - Guidance
DATOSPIR TOUCH uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
DATOSPIR TOUCH uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Harmonic emissions
EN-IEC 61000-3-2
Voltage fluctuations / Flicker emissions
EN-IEC 61000-3-3
Yes
Guidance and manufacturer’s declaration – electromagnetic immunity
DATOSPIR TOUCH is intended for use in the electromagnetic environment specified below. The costumer or the user of
DATOSPIR TOUCH should assure that it is used in such an environment.
Immunity test EN-IEC 60601 test level Compliance level Electromagnetic
Guidance environment –
Electrostatic discharge
(ESD)
EN-IEC 61000-4-2
Electrical fast transient/burst
EN-IEC 61000-4-4
Surge
EN-IEC 61000-4-5
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential
±2 kV common mode
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital environment.
To Ethernet cable.
±1 kV differential
±2 kV common mode
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines
EN-IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field
EN-IEC 61000-4-8
<5 % Ut (>95 % dip in
Ut)for 0.5 cycle
40 % Ut (60 % dip in Ut) for 5 cycles
70 % Ut (30 % dip in Ut) for 25 cycles
<95 % Ut (>5 % dip in Ut) for 5 seconds
<5 % Ut (>95 % dip in Ut)for 0.5 cycle
40 % Ut (60 % dip in
Ut)for 5 cycles
70 % Ut (30 % dip in
Ut)for 25 cycles
<95 % Ut (>5 % dip in Ut) for 5 seconds
3 A/m 3 A/m
Mains power quality should be that of a typical commercial or hospital environment. If the user of the DATOSPIR
TOUCH requires continued operation during power mains interruptions, it is recommended that the DATOSPIR
TOUCH be powered from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial of hospital environment.
NOTE Ut is the a.c. mains voltage prior to application of the test level.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 70
Guidance and manufacturer’s declaration – electromagnetic immunity
DATOSPIR TOUCH is intended for use in the electromagnetic environment specified below. The costumer or the user of
DATOSPIR TOUCH should assure that it is used in such an environment.
Immunity test EN-IEC 60601 test level Compliance level Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used no closer to any part of DATOSPIR
TOUCH, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
Conducted RF
EN-IEC 61000-
4-6
3 Vrms
150KHz to 80 MHz
3 Vrms
d
=
⎡
⎢⎣
3 .
E
5
⎤
⎥⎦
P
80 MHz to 800 MHz
Radiated RF
EN-IEC 61000-
4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d
=
⎡
⎢⎣
3 .
5
E
⎤
⎥⎦
d
=
⎡
⎢⎣
7
E
⎤
⎥⎦
P
80 MHz to 800 MHz
P
800 MHz to 2.5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked with the following symbol:
Note 1. At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which DATOSPIR TOUCH is used exceeds the applicable RF compliance level above, DATOSPIR TOUCH should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such reorienting or relocating DATOSPIR TOUCH.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 71
Recommended separation distances between portable and mobile RF communications equipment and DATOSPIR TOUCH
DATOSPIR TOUCH is intended for use in an electronic environment in which radiated RF disturbances are controlled. The costumer or the user of DATOSPIR TOUCH can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and DATOSPIR TOUCH as recommended below, according to the maximum output power of the communications equipment.
Rated
Maximum output power of transmitter
W
d
=
⎡
⎢⎣
3 .
5
3
⎤
⎥⎦
Separation distance according to frequency of transmitter
m
De 150 kHz a 80 MHz
P
De 80 MHz a 800 MHz
d
=
⎡
⎢⎣
3 .
5
3
⎤
⎥⎦
P
800 MHz to 2.5 GHz
d
=
⎡
⎢⎣
7
3
⎤
⎥⎦
P
0.01 0.12 0.12 0.23
0.1 0.37
1 1.17
0.37
1.17
0.74
2.33
10 3.69
100 11.67
3.69
11.67
7.38
23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1 At 800 MHz, the separation distance for the higher frequency applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absortion and reflection from structures, objects and people.
511-B00-MU2 • REV. 1 .02
DATOSPIR TOUCH User’s Manual 72
Annex 2.
COMPLIANCE WITH THE DATA PROTECTION
ACT. DIRECTIVE 95/46/EC
REQUIREMENTS AFFECTING THE USER OF THE EQUIPMENT
According to current legislation, the user of this equipment is
the only party responsible for saving and processing the details of his patients according to the Law.
CONFIGURATION OF DATOSPIR TOUCH PROTECTION
The DATOSPIR TOUCH spirometer has an equipment protection option that uses a PIN code, user-configurable that seeks to prevent access by unauthorized people to the equipment and, more specifically, to the private data it contains.
To comply with current legislation, users must enable this option and configure their PIN. They will be held responsible for providing this PIN to the authorized people. Thus, when the spirometer is started, the PIN will be requested and the equipment locked where this PIN is entered erroneously three times. On restarting the equipment, the unlock code (PUK), provided by the manufacturer upon purchasing the equipment, will be requested. If this code is not available, the equipment will remain locked.
• PRINTING DOCUMENTS: In the event of saving paper printouts containing patient details, these documents must be properly stored so that only duly authorized personnel have access to them. Furthermore, in the event of users deciding to dispose of the printed documents, their effective physical destruction must be ensured to avoid unauthorized access thereto.
• DATA TRANSMISSION: This spirometer can transmit files containing patient details via PC connection so that work can be subsequently carried out on them using the W20s Spirometry
Software. This software is also compliant with the Data Protection
Act, as explained in its User’s Manual.
511-B00-MU2 • REV. 1 .02