Tektronix SPG8000 Master Sync / Clock Reference Generator Instructions
Below you will find brief information for Master Sync / Clock Reference Generator SPG8000. The SPG8000 is a high-performance master sync/clock reference generator that provides accurate and reliable timing signals for video and audio applications. It features a variety of outputs, including SDI, composite, and audio, and can be used as a standalone generator or synchronized to an external reference signal. The SPG8000 is also equipped with a GPS receiver for precise time synchronization and a built-in web server for remote control and monitoring.
Advertisement
Advertisement
xx
ZZZ
SPG8000
Master Sync / Clock Reference Generator
Declassification and Security
Instructions
www.tektronix.com
*P077075000*
077-0750-00
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tektronix.com
to find contacts in your area.
Table of Contents
How to recover from clearing or removing memory from the instrument ............................. 13
SPG8000 Declassification and Security Instructions i
Table of Contents ii SPG8000 Declassification and Security Instructions
Preface
This document describes how to clear or sanitize the data storage (memory) devices in the products listed below. The instructions also describe how to declassify an instrument that is not functioning. These procedures are useful for customers with data security concerns.
Products
This document applies to the following components of the SPG8000 Master
Sync / Clock Reference Generator:
SPG8000 base generator
Option DPW (second hot-swappable redundant (backup) power supply)
Option GPS (GPS master clock synchronization)
Option BG (2 channels of composite NTSC/PAL black or HD tri-level outputs, plus 2 channels of composite NTSC/PAL test signal outputs)
Option SDI (2 channels of SD/HD SDI test signal generation on 2 outputs each)
Option 3G (3G SDI format support; Option SDI required)
Option AG (DARS output; 2 AES/EBU channels plus 4 audio tone outputs; 8
AES/EBU channels)
Required documents
To perform the procedures in this document, you will need to have access to the
SPG8000 manuals listed below. These manuals are available on the Tektronix
Web site at www.tektronix.com/downloads.
SPG8000 Quick Start User Manual
(Tektronix part number 071-3080-XX; 077-0745-XX, Japanese;
077-0746-XX, Russian)
SPG8000 Service Manual
(Tektronix part number 077-0749-XX)
SPG8000 Declassification and Security Instructions iii
Preface
Terms
The following terms may be used in this document:
Clear. This removes data on media/memory before reusing it in a secured area. All reusable memory is cleared to deny access to previously stored information by standard means of access.
Erase. This is equivalent to clear.
Media storage/data export device. Any of several devices that are used to store or export data from the instrument, such as a USB port.
Nonvolatile memory. Data is retained when the instrument is powered off.
Power off. Some instruments have a “Standby” mode, in which power is still supplied to the instrument. For clearing data, putting the instrument in
Standby mode does not qualify as powering off. For these products, you must either push a rear-panel OFF switch or remove the power source from the instrument.
Remove. This is a physical means to clear the data by removing the memory device from the instrument. Instructions are available in the product Service
Manual.
Sanitize. This eradicates the data from media/memory so that the data cannot be recovered by other means or technology. This is typically used when the device is moved (temporarily or permanently) from a secured area to a nonsecured area.
User-modifiable. The user can write to the memory device during normal instrument operation, using the instrument interface or remote control.
iv SPG8000 Declassification and Security Instructions
Clear and sanitize procedures
Memory devices
The following tables list the volatile and nonvolatile memory devices in the modules. Detailed procedures to clear or sanitize these devices are shown following the tables.
Table 1: Volatile memory devices for the SPG8000 base generator
Type and minimum size
Power PC
PPC440EPX
Function
Main processor
Type of user data
Data and settings
Backed up by battery
No
Method of modification
Data input method
Indirectly UI, remote, input signals
Location
Main / U1
User accessible
Yes
To clear
Set power on preset to factory, cycle power
MT47H64M16
64Mx16 DDR2
SDRAM
Processor memory
5M240ZTQFP144
CPLD
EP4CE6BGA256
FPGA,
270Kbits embedded memory
88E1116R gigabit ethernet
RGMII transceiver
DAC5571A
8-bit DAC
Decodes signals from the processor to generate control lines for the slots, front panel, and devices on the board.
Registers control aspects of the decoding.
Phase lock loop control,
DDS Clock sine wave gen uses some of embedded volatile memory.
10/100/1000
Ethernet
PHY control registers, integrated data FIFO buffer
Display
Contrast voltage control
Data and settings
No
Settings No
Settings No
Data No
Settings No
Indirectly
Indirectly
Indirectly
Indirectly
Indirectly
UI, remote, input signals
UI, remote, input signals
UI, remote, input signals
UI, remote
UI or remote
Main / U33,
U34
Main / U220
Main / U450
Main / U58
Main / U43
Yes
Yes
No
No
Yes
Set power on preset to factory, cycle power
Set power on preset to factory, cycle power
Set power on preset to factory, cycle power
Set power on preset to factory, cycle power
Set power on preset to factory, cycle power
To sanitize
Repeat clear process three times
Repeat clear process three times
Repeat clear process three times
Repeat clear process three times
Repeat clear process three times
Repeat clear process three times
SPG8000 Declassification and Security Instructions 1
Clear and sanitize procedures
Table 1: Volatile memory devices for the SPG8000 base generator (cont.)
Type and minimum size
5M240ZTQFP100
CPLD
Real Time
Clock +
128x8 serial
NV memory
FPGA
EP2C5F256C7N with internal registers
Function volatile keypad read-back and LED drive control registers
The RTC function is not accessible but the clock runs and will show the elapsed time since the board was built. The memory is used to store user network parameters.
Used to store instrument settings
Type of user data
Settings No
Backed up by battery
Data and settings
Yes
Settings No
Method of modification
Data input method
Indirectly UI or remote
Direct and indirect
Indirectly
UI or remote
UI
Location
Front panel /
U1
Main / U16
User accessible
Yes
Yes
GPS/Genlock/Black board, U432
No Restore factory settings
To clear
Set power on preset to factory, cycle power.
Perform the clear network settings procedure.
There is no way to clear the real time clock.
To sanitize
Repeat clear process three times.
Remove part or battery.
FPGA
EP2C35F672C6N with internal registers
Used to store instrument settings
Settings No Indirectly UI GPS/Genlock/Black board, U732
No Restore factory settings
Disconnect inputs then
Restore factory settings three times
Disconnect inputs then
Restore factory settings three times
2 SPG8000 Declassification and Security Instructions
Clear and sanitize procedures
Table 2: Nonvolatile memory devices for the SPG8000 base generator
Type and minimum size Function
Serial EEPROM
AT24C01B 128x8
Processor boot configuration data
Type of user information stored
Method of modification
Data input method
None None Factory
JTAG
Location
Main / U21
NOR Flash
PC28F512P30
64Mx16
NAND Flash
K9F1G08U0A
128Mx8
Serial MRAM
MR25H256
32Kx8
CPLD,
5M240ZTQFP144,
8192 bits flash
Serial EEPROM
AT24C01B 128x8
Main storage for processor.
Contains
Operating system, application programs,
MAC address, instrument family type, FPGA programming files, signal, logo, text overlay and picture files.
Not used.
Oven calibration data
Decodes signals from the processor to generate control lines for the slots, front panel, and devices on the board. NV flash is not used
Run time data for power supply showing time of use and temperature / time related use
Both data and settings
None
Settings
Settings
Data
Directly and indirectly
Presets are saved from
UI, files and presets are saved from USB or ethernet.
Main / U101, U102 Yes
None
Indirectly
Indirectly
Indirectly
None
Perform calibration
Software upgrade
Automatic over time
Main / U103
Main / U22
Main / U220
User
Accessible
No
No
Yes
No
U1 on Power
Supply module; also on Option DPW secondary Power
Supply module
Yes
To clear
N/A
Perform the delete presets and user files from memory procedure.
To sanitize
Remove part from board and destroy.
Remove part from board and destroy.
N/A
Perform calibration
N/A
None
N/A
Remove part from board and destroy.
N/A
N/A
Remove parts from board and destroy; can also remove supply from instrument and leave it in a secured area; replace supply with one that has not been in a secure area
N/A CPLD
5M240ZTQFP100,
8192 bits flash memory not used
Interfaces with signals from the main PLD, drives LEDs and reads the buttons.
nonvolatile flash memory is not used.
Settings Indirectly Software upgrade
Front panel / U1 No
SPG8000 Declassification and Security Instructions 3
Clear and sanitize procedures
Table 2: Nonvolatile memory devices for the SPG8000 base generator (cont.)
Type and minimum size Function
Flash memory
4 M X 8
Configuration of
FPGA
Type of user information stored
Method of modification
Data input method
None None Written by processor system from
Main board
Location
GPS/Genlock/Black board, U0331
User
Accessible
No
Flash memory
4 M X 8
Flash memory
4 M X 8
Calibration for the module
Stores leap second information
Settings
Unused None
The values are a function of how the user calibrates the module and when it was last receiving the
GPS signal
Stored during calibration
Stores leap second information obtained by satellite
GPS/Genlock/Black board, U0332
Yes
None None GPS/Genlock/Black board, U0741
No
To clear
None
Manually calibrate the module back to the factory default values
Allow the module to receive satellite information for
30 minutes to update the leap second field
None None
To sanitize
N/A
No customer information is stored in this flash memory
Remove and destroy the memory device or destroy the entire board
Table 3: Nonvolatile memory devices for Option GPS
Type and minimum size
GPS receiver assembly
Function
Calculates timing and position information
May contain user data 1
Yes if operated in fixed mode
Data input method
Location is stored when operated in fixed mode
1 During normal instrument operation.
Location
Mounts above the main module board and connects to J12 with a ribbon cable, and to the antenna connector on the rear panel
To clear
Save a position from a location different from the one that was last saved. This will erase the previous position and save the new one
To sanitize
Remove and destroy the module or destroy the entire board
4 SPG8000 Declassification and Security Instructions
Clear and sanitize procedures
Table 4: Volatile memory devices for Option BG
Type and minimum size
PLD
EPM7128A with internal registers
Function
Used to store instrument settings
Type of user data
Settings No
Backed up by battery
Method of modification
Data input method
Indirectly UI
Location
Option BG
Black board,
U300
User accessible
No
To clear
Restore factory settings
FPGA
EPM1K50 with internal registers
Used to store instrument settings and to generate test signals
Settings No
SRAM 128K ×
36 IDT71V546
Used to store
Composite test signals
Signals No
Indirectly UI Option BG
Black board,
U400 U500
U600 U700
No
Indirectly UI selection of format
Option BG
Black board,
U620 U720
No
Restore factory settings
Restore factory settings
To sanitize
Disconnect inputs then
Restore factory settings three times
Disconnect inputs then
Restore factory settings three times
Disconnect inputs then
Restore factory settings three times
SPG8000 Declassification and Security Instructions 5
Clear and sanitize procedures
Table 5: Volatile memory devices for Option AG
Type and minimum size
PLD
EPM7128A with internal registers
Function
Used to store instrument settings
Type of user data
Settings No
Backed up by battery
Method of modification
Data input method
Indirectly UI
Location
Option AG
Audio board,
U200
User accessible
No
To clear
Restore factory settings
FPGA
EPF10K30 with internal registers
Used to store instrument settings and to generate test signals
Settings No
SRAM 512K ×
8 CY7C1049
Used to store audio sine wave signals
Signals No
Indirectly UI Option AG
Audio board,
U300
No
Indirectly UI selection of signal
Option AG
Audio board,
U310
No
Restore factory settings
Restore factory settings
To sanitize
Disconnect inputs then
Restore factory settings three times
Disconnect inputs then
Restore factory settings three times
Disconnect inputs then
Restore factory settings three times
6 SPG8000 Declassification and Security Instructions
Clear and sanitize procedures
Table 6: Nonvolatile memory devices for Option SDI
Type and minimum size
Flash 32 M x 16
Function
FPGA configuration, unique module
ID, option key
May contain user data 1
Yes
PLD with internal
512 x16 byte user flash memory
DDR2 32 M x 16
PLD addresses decoding, memory not used
Used for frame picture and logo overlay functions
No
Yes
Data input method
Written by processor system on
SPG8000 main board using
PLD U226.
Option 3G only: User can enter option key string from
UI. String must be correct to enable Option
3G.
Programmed at factory
SRAM 256 K x 16 Used for text and circle overlay functions
Yes
Yes,
Indirectly
Load frame picture and or logo BMP into mainframe, then select frame picture or logo overlay function
Load text file to mainframe or enter text string from keyboard
Location
389-4093-00 board, U321
389-4093-00 board, U226
389-4093-00 board,
U618, U711
389-4093-00 board
U521, U0521,
U621, U0621,
U622, U0622
Calibrate output levels
389-4485-00 output board
U1 and U5
To clear
Overwrite the option string with dummy characters.
None
Remove power
If text is saved in power on preset, then should restore factory preset, save as power on preset, then cycle power three times
Store calibration level at mid-scale of 50
Dual 2048 step digital potentiometer with
EEPROM with the two parts in parallel, limited in software to
100 steps of range
FPGA with 90 K
Logic elements and 4.5 Mb of
RAM
Holds calibration data
Generates signals from logic and memory; also used for burn-in timecode
Yes, indirectly by signal selection and instrument configuration
No
Programed from flash at bootup, changes state as instrument operates
None
389-4093-00 board, U811
389-4093-00 board, U224
Remove power
None 128 K x 8
EEPROM
1
Unused
During normal instrument operation.
To sanitize
Remove part from board and destroy
Remove part from board and destroy
Remove part from board and destroy
Remove part from board and destroy
Store calibration level at min and max three times then store at mid scale of 50
Set instrument to power up on factory default and cycle power three times
Remove part and destroy
SPG8000 Declassification and Security Instructions 7
Clear and sanitize procedures
To restore the factory default settings
Restore the instrument to the factory default settings as follows:
1. Press the SYSTEM button until SYSTEM: PRESET appears.
2. Press the ENTER button. You will see SYSTEM : PRESET : RECALL displayed.
3. Press the left (◄) or right (►) arrow button until Factory Default appears.
4. Press the ENTER button.
5. Press the BACK button to exit the Preset menu.
To set the factory default settings as the power on state
1. Perform the previous procedure to restore the factory preset/default state.
2. Press the SYSTEM button until SYSTEM : PRESET appears.
3. Press the ENTER button. You will see SYSTEM : PRESET : RECALL displayed.
4. Press the up (▲) or down (▼) arrow button to select SYSTEM : PRESET
: SAVE
5. Press the left (◄) or right (►) arrow button until the display shows SYSTEM
: PRESET : SAVE Power On Default.
6. Press the ENTER button. The display will flash as it saves the preset.
7. Press the BACK button to exit the Preset menu.
To clear the network settings
1. Press the SYSTEM button until SYSTEM : PRESET appears.
2. Press the up (▲) or down (▼) arrow button to select SYSTEM :
NETWORK.
3. Press the left (◄) or right (►) arrow button to select Setup, and then press the ENTER button.
4. Press the up (▲) or down (▼) arrow button to select SYSTEM : NET
SETUP : DHCP.
5. Press the left (◄) or right (►) arrow button to select Disable, and then press the ENTER button.
NOTE.
The DHCP feature must be disabled before you can clear the network settings stored in the instrument memory.
8 SPG8000 Declassification and Security Instructions
Clear and sanitize procedures
6. Press the up (▲) or down (▼) arrow button to select SYSTEM : NET
SETUP : IP ADDRESS, and then press the ENTER button.
7. Use the left (◄) or right (►) arrow buttons to select each number in the IP address, and use the up (▲) or down (▼) arrow buttons to change the address values to a safe default number such as all zeros. Press the ENTER button to save the changes.
8. Repeat steps 6 and 7 for the Subnet Mask and Gateway addresses.
9. Press the BACK button to exit the NETWORK submenu.
To delete presets and user files from memory
Backup existing presets or user files. If you want to backup your presets or user files before you delete them so that you can restore them at a later time, perform
the following steps. Otherwise, proceed to step 12.
1. Connect a USB drive to the SPG8000 generator.
2. Press the SYSTEM button to select SYSTEM : PRESET.
3. Verify the status of the USB drive:
a. Press the up (▲) or down (▼) arrow button to select SYSTEM : USB
(status).
b. Verify that the USB status is Mounted.
c. If the status is Not Mounted, press the ENTER button to mount the USB drive.
4. Press the up (▲) or down (▼) arrow button to select SYSTEM :
BACKUP/RESTORE.
5. Press the left (◄) or right (►) arrow button to select one of the following:
Backup Presets to USB: Backs up only the instrument presets to the USB drive. If you are backing up only presets, you will need up to 1 MB of free space on the USB drive.
Backup All User Data to USB: Backs up all user data to the USB drive.
When all user data is backed up, every user file except for the FPGA configuration files will be copied to the USB drive. User files are defined as any file that exists in the /app/F0 directory on the instrument. The files in this directory can include presets, signal files, frame picture files, logo files, sequence files, and for Option SDI and Option 3G, can also include text and font files. The user data files do not include the application software or any Linux files.
NOTE.
If you are backing up all of the user data in the instrument, you will need up to 96 MB of free space on the USB drive.
SPG8000 Declassification and Security Instructions 9
Clear and sanitize procedures
6. Press the ENTER button to start the selected backup operation.
7. A warning message is displayed asking you to verify the operation. Press the
ENTER button to proceed or press the BACK button to cancel the operation.
8. Monitor the progress of the file copy operation:
a. During the first stage of the copy process, the display shows the names of the files as they are copied. Depending on the size of the files being copied, this may take a couple of minutes.
b. When all of the files have been copied, the message “Finishing copy...” is displayed while the file system buffers are emptied. This message may appear for a couple of minutes.
c. When the backup operation is complete, the message “Backup completed successfully” is displayed before the menu display returns to normal.
NOTE.
If the USB drive runs out of memory before the backup operation is complete, the operation will copy as much data as it can and will then display an error message stating that the backup operation was incomplete because the drive or device is full. Press the ENTER or BACK button to clear the error message.
9. After the backup operation is complete, press the up (▲) or down (▼) arrow button to select SYSTEM : USB (Mounted).
10. Press the ENTER button to unmount the USB drive. After the message “You may now safely remove the USB drive” is displayed, the menu readout should change to SYSTEM : USB (Not mounted).
11. Remove the USB drive from the instrument.
Delete the presets and user files from the instrument.
12. Connect the instrument and a computer using an Ethernet connection as described in the SPG8000 Quick Start User Manual.
13. On the computer, open an FTP client such as Windows Explorer.
14. In the Windows Explorer command box, enter “ftp://IP address”, where IP address is the network address assigned to the TG8000 generator.
NOTE.
Some FTP clients may ask for login credentials. In this case, login with a user name of “anonymous” and no password.
10 SPG8000 Declassification and Security Instructions
Clear and sanitize procedures
15. The Windows Explorer window should immediately connect to the instrument and show the directory structure as shown below.
Figure 1: Example FTP view of the instrument file structure
16. In the CPU directory, delete any preset files.
17. In the SEQUENCE directory, delete any sequence files.
18. In each of the module directories, delete all of the preset files and any user-created picture, logo, text, or signal files.
NOTE.
In addition to the user-created files, the module directories will also contain the factory default files for the standard signals and logos that are supplied with the instrument. You do not need to delete the factory default files.
19. After you have deleted the desired files, close the FTP window.
SPG8000 Declassification and Security Instructions 11
Clear and sanitize procedures
12 SPG8000 Declassification and Security Instructions
Troubleshooting
How to clear or sanitize a nonfunctional instrument
To sanitize a nonfunctional instrument, remove the Main board,
GPS/Genlock/Black board, SDI board, and all Power Supply modules, and then return the instrument to Tektronix for installation of new boards and modules.
How to recover from clearing or removing memory from the instrument
The procedure to recover from clearing or removing memory depends on whether the memory was on a generator card or on the instrument mainframe. Perform the appropriate procedure as described in this section.
To recover from clearing or removing memory from a generator card
Perform the following procedure to recover a generator card:
1. Power on the instrument.
2. Calibrate the generator card as instructed in the SPG8000 Service Manual.
3. Reconfigure the generator card for your specific application. You can find configuration information in the SPG8000 Quick Start User Manual.
4. For Option GPS only:
a. Connect a GPS signal to the generator card and let it run for 30 minutes.
This allows the leap second and position information to be reacquired.
b. You are finished with the Option GPS recovery; the GPS generator card will reload the system software on start up.
5. For Option SDI only, verify that Option 3G is enabled:
NOTE.
You need to perform the following steps only if the instrument had
Option 3G enabled prior to being sanitized.
a. Press the SYSTEM button.
b. Press the up (▲) or down (▼) arrow button to select SYSTEM :
OPTIONS.
c. The second line of the display lists the enabled options. Check that 3G is displayed.
e. If Option 3G is not enabled, you must reenter the option key for Option 3G as described in the SPG8000 Quick Start User Manual.
SPG8000 Declassification and Security Instructions 13
Troubleshooting
To recover from clearing or removing memory from the
SPG8000 base instrument
Perform the following procedure to recover the SPG8000 base instrument:
1. Power on the instrument.
2. Calibrate the mainframe oscillator oven as instructed in the SPG8000 Quick
Start User Manual. To calibrate the oscillator, you can use a GPS signal if
Option GPS is installed. Otherwise, you can use a 10 MHz signal or a video reference signal that is known to be at a precise frequency.
3. Configure the network parameters for the instrument as instructed in the
SPG8000 Quick Start User Manual.
4. If you backed up the presets and user data files onto a USB drive before you removed the mainframe memory, restore the presets or user data files as described below:
a. Connect the USB drive to the SPG8000 generator that you used to back up the presets or user data files.
b. Press the SYSTEM button to select SYSTEM : PRESET.
c. Press the up (▲) or down (▼) arrow button to select SYSTEM : USB
(status).
d. Verify that the USB status is Mounted.
e. If the status is Not Mounted, press the ENTER button to mount the USB drive.
f. Press the up (▲) or down (▼) arrow button to select SYSTEM :
BACKUP/RESTORE.
g. Press the left (◄) or right (►) arrow button to select Restore Presets
from USB or Restore All User Data from USB.
h. Press the ENTER button to start the selected restore operation.
i.
A warning message is displayed asking you to verify the operation. Press the ENTER button to proceed or press the BACK button to cancel the operation.
j. During the first stage of the copy process, the display shows the names of the files as they are copied. Depending on the size of the files being copied, this may take a couple of minutes.
k. When all of the files have been copied, the message “Finishing copy...” is displayed while the file system buffers are emptied. This message may appear for a couple of minutes.
l.
When the restore operation is complete, the message “Restore completed successfully” is displayed before the menu display returns to normal.
14 SPG8000 Declassification and Security Instructions
Troubleshooting
m. After the restore operation is complete, press the up (▲) or down (▼) arrow button to select SYSTEM : USB (Mounted).
n. Press the ENTER button to unmount the USB drive. After the message
“You may now safely remove the USB drive” is displayed, the menu readout should change to SYSTEM : USB (Not mounted).
o. Remove the USB drive from the instrument.
5. If you did not restore the presets or user data files from a USB drive in step 4,
then adjust the display contrast as follows:
a. Press the SYSTEM button to select SYSTEM : PRESET.
b. Press the up (▲) or down (▼) arrow button to select SYSTEM : LCD
CONTRAST.
c. Press the left (◄) or right (►) arrow button to set the display contrast to the desired level.
NOTE.
Perform the following steps if you want to save the display contrast setting so that it will be restored when the instrument is powered on.
d. Press the SYSTEM button to select SYSTEM : PRESET, and then press the ENTER button.
e. Press the up (▲) or down (▼) arrow button to select SYSTEM :
PRESET : SAVE.
f. Press the left (◄) or right (►) arrow button to select Power On Default.
g. Press the ENTER button to save the current instrument settings as the power on default.
SPG8000 Declassification and Security Instructions 15
Download
Advertisement
Key features
High-performance master sync/clock reference generator
Variety of outputs (SDI, composite, audio)
Standalone operation or external reference synchronization
Built-in GPS receiver for precise time synchronization
Built-in web server for remote control and monitoring
Frequently asked questions
The SPG8000 offers a range of outputs, including SDI, composite, and audio. These outputs can be configured to generate various timing signals, such as black burst, color bars, and test patterns.
Yes, the SPG8000 can be synchronized to an external reference signal, such as a 10 MHz signal or a video reference signal. This allows for accurate timing synchronization across multiple devices in a system.
The document provides detailed instructions on how to clear or sanitize the memory on the SPG8000. These procedures involve removing or erasing data from various memory devices within the instrument.
The document also includes recovery procedures that involve calibrating the instrument, configuring network parameters, and restoring any backed-up presets or user data files.