Non-contact Infrared Thermometer

User Manual

Please read the manual carefully before using this product.

A: Model; B: Page Up; C:Page Down ; D: Detection port; E: Display; F: ON/Scan;

G: Battery Holder.

TABLE OF CONTENTS

1. Instruction……………………………………………………………………………………………………………………………1

2. Basic Principle………………………………………………………………………………………………………………………1

3. Feature…………………………………………………………………………………………………………………………………1

4. Technical Specifications…………………………………………………………………………………………………………2

5. Parts………………………………………………………………………………………………………………………………………3

6. Replacing Battery…………………………………………………………………………………………………………………3

7. Setting Up Your Non-contact Infrared Thermometer………………………………………........3

8. Taking a Body Temperature Measurement……………………………………………………………………………3

9. Taking a Object Surface Temperature Measurement……………………………………………………………..4

10. Viewing Data Memory…………………………………………………………………………………………………………4

11. Problems & Troubleshooting……………………………………………………………………………………………….4

12. Product Maintenance & Attention……………………………………………………………………………………….5

13. EMC Declaration…………………………………………………………………………………………………………………..6

14. Quality commitment and after-sales service……………………………………………..………………………..12

一、 Instruction

This product measures human body temperature by collecting infrared thermal radiation from human forehead. User only needs to point the probe at the forehead and press ON/SCAN

Button to measure the body temperature. It widely used in schools, customs, hospitals, homes.

It is a second-class medical device, which belongs to the internal power supply equipment, type B application part, and the liquid-proof level is ordinary equipment. It cannot be used in the presence of flammable anesthetic gas and air, oxygen or nitrous oxide.

Scope of application: display the body temperature of the measured object by measuring the heat radiation of the forehead

The meanings of the legend and warning mark as below:

Contraindications: N/A

二、 Basic Principle

Any object will release infrared radiation energy, and its surface temperature directly determines the size and wavelength of the radiation energy. Based on this principle, this product uses a high-precision infrared sensor that is specifically used to detect the human body emitting infrared radiation with a wavelength of 5-14um, and accurately measures the human body temperature through accurate calculations and various compensation corrections.

三、 Feature

1. Using high-precision infrared sensor;

2. Strong adaptability to ambient temperature;

3. Automatically save the last measured value;

4. Large LCD screen with high brightness backlight;

5. Two temperature units of Celsius and Fahrenheit can be selected;

6. Automatic shutdown to save power and energy;

7. Clean and hygienic: non-contact measuring forehead temperature, measuring distance 3-5cm, not touching human skin, preventing cross infection.

8. One-click measurement, one-click check memory;

9. Multiple sets of memory: It can store 30 sets of measurement data for analysis and comparison.

四、 Technical Specifications

Model

Measurement Method

Measuring Distance

Measuring range

Body temperature mode

Object surface mode

Maximum allowable error

Body temperature mode

Object surface mode

Display resolution

Operating environment

Transportation storage environment

Power

Battery reminder

Backlight

Display unit

Automatic shut-down

Software release version

HZK-801, HZK-802, HZK-803, HZK-805, HZK-806,

HZK-807, HZK-808, HZK-809, HZK-810

Non-contact

3cm-5cm

32.0

℃ -42.0

25.0

℃ -45.0

34.0

℃ -34.9

35.0

℃ -42.0

25.0

℃ -34.9

35.0

℃ -42.0

土 0.3

土 0.2

土 1.0

土 0.3

1.0

Temperature: 16 ℃ -35 ℃

Humidity.<=85%

Temperature 20 ℃ -55 ℃ ,Humidity ≤93%

D.c.3V(2 × AAA)

Low bdttery alert

High brightness backlight

Centigrade (℃) / Fahrenheit (º F )

Less than 20 seconds

V1

五、 Parts

This product is mainly composed of display screen, infrared detection head, PCBA board and housing.

List of accessories: 1 instruction manual (including warranty card), 1 certificate, 2 AAA batteries.

六、 Replacing Battery

1) Push the battery door downwards and backwards firmly, insert 2 AA batteries, and cover the battery cover. When installing, please carefully check the positive and negative signs on the battery. Pay attention that the positive and negative electrodes of the battery cannot be reversed.

2) If not used for a long time (more than 3 months), please remove the battery.

七、 Setting Up Your Non-contact Infrared Thermometer

In the power-on state, click the mode button to switch modes.

八、 Taking a Body Temperature Measurement

This product provides you with a method of measuring the body temperature of the frontal temperature, but it cannot be used as a substitute for a doctor's diagnosis and treatment. In addition, the individual's body temperature is different. Please use the memory function to understand your daily temperature and use it as a reference for whether the temperature rises.

1) Point the thermometer at the center of the forehead-above the center of the eyebrow and keep it vertical. The measuring part must not be covered by hair.

2) Press the key to start

Note 1: The body temperature mode is the preset mode.

Note 2: The thermometer will be performed automatically after power on

3) About 2 seconds, the thermometer screen displays "body temperature" and displays the measurement result.

4) Automatic shutdown: Automatic shutdown after 15 seconds without operation.

Before measuring, please make sure that the subject has not bathed or exercised within 30 minutes, and has been kept in a stable environment for at least 5 minutes. Three measurements are recommended. If the three measurements are different, choose the highest temperature value.

九、 Taking a Object Surface Temperature Measurement

This product is a function for you to measure the surface temperature of an object.

1) When the instrument is in the measurement-on state, press the Mode key. The LCD displays the "Surface Temperature" character, indicating that it is set to the object temperature measurement mode.

2) Automatic shutdown: Automatic shutdown after no operation for about 15 seconds. When the thermometer is taken out from a place with a large difference in the ambient temperature to be measured, the thermometer should be placed in a new environment for at least 30 minutes before measuring.

十、 Viewing Data Memory

In the on or off state, press the MEMO key to enter the memory view.

1) The larger the memory number, the earlier the measured value, and the smaller the value, the closer the measured value.

2) If there is no memory value, “————” is displayed.

3) The thermometer can store up to 30 groups of recent measured values, and the values of the remaining 30 groups are automatically overwritten in chronological order.

十一、 Problems & Troubleshooting

Display/Problems Express meaning

Display: Hi

Display: Lo

Display: POS

Display: Lo

Display: POS

Environmental temperature too high

Environmental temperature too low

Battery level display

Possible Reasons

1.Body temperatuer mode,display ℃ temperatuer higher than 42 ℃

2.Object temperature mode,display temperature is higher than 42 ℃

1.Body temperatuer mode,display temperatuer is below 34 ℃

2.Object temperature mode,display temperature is below 25 ℃

Low battery,please replace with new battery immediately

十二、 Product Maintenance & Attention

Please keep the inner cavity of the sensor and probe clean, otherwise it will affect the measurement accuracy.

1. Cleaning method:

1) Surface cleaning: Wipe the dirt with a clean soft cloth or cotton swab moistened with a little medical alcohol or water.

2) Clean the inner cavity of the sensor and probe: Wipe the inner cavity of the probe or the top of the sensor gently with a clean soft cloth or cotton swab with a little medical alcohol. Do not use until the alcohol has completely evaporated.

2. Note:

1) Before use, please read this instruction manual carefully and make sure the battery is installed.

2) It is forbidden to immerse the thermometer in any liquid, and it is forbidden to be placed in an excessively high or low temperature environment for a long time.

3) It is forbidden to collide, drop or mix with sharp objects, and it is forbidden to disassemble by yourself.

4) The thermometer cannot be used in the sun or water.

5) Do not use in a strong electromagnetic interference environment.

6) Keep the thermometer out of the reach of children.

7) Do not open the battery door during the test to avoid touching the live parts and the patient at the same time.

8) Do not touch the lens inside the gun head with your fingers.

9) Sweat stains on the forehead, coverings such as hair, hats, or scarves may cause the measurement temperature to be too low, so please use it correctly to ensure the correct measurement results

10) It is recommended to practice more to become familiar with the measurement method, and try not to change the product factory settings. Measurement results are not a substitute for physician diagnosis.

11) No special maintenance is required during use, please contact the seller or manufacturer if any trouble occurs.

12) Please dispose of waste and residues at the end of the service life of the product according to local laws and regulations.

十三、 EMC Declaration

This product complies with the EMC (Electromagnetic Compatibility) standard YY0505-2012 for the safe use of medical electrical equipment. The EMC standard is a standard developed for the safe use of medical electrical equipment. Electromagnetic interference from other devices

(mobile phones, etc.) is controlled within a certain range. YY0505-2012 (item 5.5.1.1) provides detailed information about the EMC environment required to provide users with safe equipment operation. The following is a description of EMC-related technical descriptions.

When this product works in the electromagnetic environment locked in this EMC technical data, the basic performance described in its scope of application will not be affected.

Definition of EMC (Electromagnetic Compatibility);

EMC (Electromagnetic Compatibility) refers to the ability to meet the following two requirements.

Capacitive electromagnetic interference noise will not be emitted to other nearby electronic equipment. (radiation)

The device can function normally in an electromagnetic environment in which other electronic devices emit noise or the like. (Immunity)

EMC (Electromagnetic Compatibility) related technical descriptions;

Note: The basic performance is: "During the EMC test, the thermometer display normally displays, continuous temperature measurement, and trouble-free operation."

Caution: Portable and mobile RF communication equipment may affect the normal operation of this instrument.

Warning: It is the responsibility of the user to ensure the electromagnetic environment of the equipment, so that the instrument can work normally.

Warning: Do not use this device near strong radiation sources, otherwise it may interfere with the normal operation of this instrument.

Caution: Except for transducers, accessories, and cables sold by the manufacturer of this instrument or system as spare parts for internal components, the use of unspecified transducers, accessories, and cables may result in increased emissions or immunity of the equipment or system The reduction.

Warning: This instrument should not be used close to or stacked with other equipment. If it must be used close to or stacked, it should be observed to verify that it can operate normally in its used configuration.

It is recommended to use d.c.3V (2 AAA batteries) provided by the manufacturer

Table 1-Guidelines and manufacturer's declaration — Electromagnetic radiation —

Guidance and manufacturer's declaration-electromagnetic emissions

This infrared thermometer is intended for use in the electromagnetic environment specified below, and the purchaser or user should guarantee its use in this electromagnetic environment:

Launch test Compliance

RF emission

GB 4824

Team 1

Electromagnetic Environment-Guide

This infrared thermometer uses radio frequency energy only for its internal functions. As a result, its

RF emissions are low and the possibility of interference with nearby electronic equipment is extremely low.

RF emission

GB 4824

Harmonic radiation

GB17625.1

Voltage fluctuation / flicker emission

GB17625.2

Class B

Not applicable

Not applicable

This product is suitable for use in all facilities.

Table 2—Guidelines and manufacturer's declaration—Electromagnetic immunity—

Guidance and manufacturer's declaration-electromagnetic immunity

This infrared thermometer is intended for use in the electromagnetic environment specified below, and the purchaser or user should guarantee its use in this electromagnetic environment.

Immunity test IEC 60601 test level Coincidence level Electromagnetic

Environment-Guide

Electrostatic discharge GB /

T17626.2

± 6kV contact discharge

± 8kV air discharge

± 6kV contact discharge

± 8kV air discharge

The floor is recommended to be wood, concrete or tile floor. When the floor is coated with synthetic material, the relative humidity is recommended to be at least 30%.

Electrical fast transient pulse group GB /

T17626.4

± 2kV power cord

± 1kV input / output line

Not applicable Not applicable

Surge GB /

T17626.5

Not applicable Not applicable

Voltage sags on power input lines, short-term interruptions and voltage changes GB /

T17626.11

± 1kV wire-to-wire

± 2kV line to ground

<5% U

T for 0.5

weeks (on U

T

,> 95% sag)

40% U

T for 5 weeks

(on U

T

, 60% sag)

70% U

T for 25 weeks

(on U

T

, 30% slump)

<5% U

T for 5 seconds (on U

T

,>

95% sag)

Not applicable Not applicable

Power frequency magnetic field

(50 / 60HZ) magnetic field

GB / T17626.8

3A/m 3A/m, 50H

Z

If the instrument does not work properly, it is necessary to keep the instrument away from the power frequency magnetic field source.

Note: U

T is the AC network before the test voltage is applied.

Table 3—Guidelines and manufacturer declarations—Electromagnetic immunity—

Guidance and manufacturer's declaration-electromagnetic immunity

This infrared thermometer is intended for use in the electromagnetic environment specified below, and the purchaser or user should guarantee its use in this electromagnetic environment.

Immunity test IEC 60601 test level

Coincidence level Electromagnetic Environment-Guide

RF conduction GB

/ T17626.6

RF radiation GB /

T 17626.3

3V (effective value)

150KH

Z

-80MH

Z

3V/m

80MH

Z-

2.5GH

Z

Not applicable

3V/m

Portable and mobile RF communications equipment should not be used closer to any part of this product, including cables, than the recommended isolation distance. This distance is calculated by the transmitter's frequency response company.

Recommended separation distance

Among them, p is the maximum rated output power of the transmitter specified by the transmitter manufacturer, in watts (W), d is the recommended separation distance, in meters (m). The electric field strength of the fixed RF transmitter is determined by the on-site survey (a) of the electromagnetic field, and should be lower than the compliance level in each frequency band range.

Interference may occur near the equipment marked with the following symbol.

Note 1: At the frequency of 80MHZ and 800MHZ, the formula of the higher frequency band is used.

NOTE 2: These guidelines may not be suitable for all situations. Electromagnetic propagation is affected by the respiration and reflection of buildings, objects, and the human body.

a) Stationary transmitters, such as: base stations for wireless (cellular / fly-free) telephones and ground mobile radios, amateur radios, FM and FM radio broadcasts, and television broadcasts, the theoretical field strengths cannot be accurately predicted in theory. To assess the electromagnetic environment of fixed RF transmitters, surveys of electromagnetic fields should be considered. If the measured field strength of the electronic thermometer is higher than the applicable RF compliance level above, the electronic thermometer should be observed to verify that it can operate normally. If abnormal performance is observed, supplementary measures may be necessary, such as reorienting or positioning the electronic thermometer b) In the frequency range of 150KHZ-80MHZ, the magnetic field strength should be lower than 3V / m.

Table 4—Recommended separation distances between portable and mobile RF communications equipment and this product

Recommended separation distance between portable and mobile RF communication equipment and this infrared thermometer

This infrared thermometer is expected to be used in an electromagnetic environment where radio frequency radiation disturbance is controlled. Depending on the maximum rated output power of the communication equipment, the purchaser or user can maintain the portable and mobile RF communication equipment (transmitter) and the infrared thermometer as recommended below. Minimum distance between them to place electromagnetic interference.

Transmitter's rated maximum output power (W)

0.01

0.1

1

10

Not applicable

Not applicable

Not applicable

Not applicable

0.12

0.38

1.2

3.8

0.23

0.73

2.3

7.3

100 Not applicable 12 23

For the maximum rated output power of transmitters not listed in the table above, the recommended separation distance d is in meters (m), which can be determined by the formula in the corresponding transmitter frequency column, where p is the emission provided by the transmitter manufacturer Maximum rated output power of the unit, in watts (W).

Note 1: At the frequency of 80MHZ and 800MHZ, the formula of higher frequency range is adopted.

Note 2: These guidelines may not be suitable for all situations. Electromagnetic propagation is affected by absorption and reflection from buildings, objects, and the human body.

Warranty Card

Cardholder

Address

Model

Quality warranty:

Contact Number

Date of manufacture

1) The life of this product is 5 years from the date of purchase. With a shopping invoice, you can enjoy a one-year free warranty and lifetime maintenance.

2) The battery and packaging are not covered by the warranty.

Time Cardholder

Note:

1: If your contact information has changed, please notify us in time!

2: The replaced parts belong to our company;

3: Consulting Tel:

4: Dear customer, please fill in the (After-sales Service Card) truthfully and send it back in order to get more and better after-sales service.

Quality commitment and after-sales service

1. The period of use of this product is five years from the date of purchase. With a shopping invoice, you can enjoy a one-year free warranty and lifetime maintenance.

2. The battery and packaging are not covered by the warranty.

3. For the following damage caused by the user, please forgive us for not being able to provide free warranty service: a) Failure caused by unauthorized disassembly and modification of the product; b) Failure caused by accidental drop during use or handling; c) failure caused by failure to follow the correct instructions in the instructions; d) failure caused by lack of reasonable maintenance; e) When requesting free warranty service, you must hold a warranty card with the date of purchase and the seal of the purchase dealer (including the dealer name and address). Be sure to ask your dealer to sign the warranty card when purchasing this product.

f) Repair services outside the warranty will be charged according to the corresponding regulations; g) When requesting a free warranty service, please take this product to the company's distribution points for repair.

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Key Features

  • Non-contact measurement for hygiene and safety
  • High-precision infrared sensor for accurate readings
  • Body and object surface temperature measurement modes
  • Large LCD screen with high brightness backlight
  • Automatic shutdown to conserve battery power
  • Multiple memory sets for data storage and comparison

Related manuals

Frequently Answers and Questions

What is the measurement range of the thermometer?
Body temperature mode: 32.0℃-42.0℃; Object surface mode: 25.0℃-45.0℃
What is the maximum allowable error of the thermometer?
Body temperature mode: ±0.3℃ (34.0℃-34.9℃), ±0.2℃ (35.0℃-42.0℃); Object surface mode: ±1.0℃ (25.0℃-34.9℃), ±0.3℃ (35.0℃-42.0℃)
How many sets of measurement data can the thermometer store?
30 sets

advertisement